WorldWideScience

Sample records for blood gas monitoring transcutaneous

  1. Transcutaneous monitoring of blood gases: is it comparable with arterialized earlobe sampling?

    Science.gov (United States)

    Dawson, S; Cave, C; Pavord, I; Potter, J F

    1998-03-01

    Researchers are increasingly looking for reliable non-invasive methods of assessing blood gas concentrations, and several new techniques have recently become available. Values derived using arterialized earlobe samples have been found to be comparable with conventional arterial samples, and recent studies have compared transcutaneous blood gas analysis with the traditional arterial samples and found a reasonable level of agreement in particular for the partial pressure of carbon dioxide. There are no data comparing oxygen and carbon dioxide partial pressures (pO2, pCO2) derived from arterialized samples with one of the newer transcutaneous techniques. We therefore simultaneously studied arterialized earlobe blood gas samples and values for pO2 and pCO2 obtained by a transcutaneous monitor (TINA, Radiometer, Copenhagen) in 26 subjects with varying blood gas values. There was a close agreement between the two methods for assessment of pCO2 [mean difference (95% C.I.) between transcutaneous and earlobe values 0.25 kPa (-0.004, 0.5 kPa)], but not for pO2 [1.71 kPa (0.35, 3.07 kPa)]. Similarly, the limits of agreement were narrow for pCO2 compared to those for pO2 (-0.98, 1.47 kPa and -6.44, 3.02 kPa respectively). We conclude that transcutaneous measurement of pCO2 using the TINA is acceptable in the research setting, whereas assessment of pO2 cannot reliably be made using this technique. PMID:9692127

  2. Optical Transcutaneous pCO2 Sensor using Soft Lithography Method for Arterial Blood Gas Analysis

    Science.gov (United States)

    Kang, Byoung-Ho; Kim, Do-Eok; Leem, Myoung-Kun; Kwon, Dae-Hyuk; Lee, Kwang-Man; Kang, Shin-Won

    2008-10-01

    In this study, we carried out development of non-invasive optical transcutaneous partial pressure of carbon dioxide, pCO2, monitoring system. The purpose of this system is to detect CO2 from outer skin, not from the arterial blood-gathering method. There are advantages about a reduction of analysis time and real-time monitoring that this system might be available. The measurement system is composed of the IR lamp, the pyroelectric sensor including a 4.26 μm optical filter, the optical gas reaction chamber and the signal processing circuit. The optical reaction length of chamber was reduced by 1 mm using the soft-lithography method which CO2 is exhausted in human body as we considered. The fabricated pCO2 monitoring system showed linear result of 6.50×10-6 absorbance/ppm sensitivity for CO2 concentration from 0 ˜5,000 ppm by MFC and about 2 seconds of fast response time. The proposed system can be used in the optical biosensor field for the medical diagnosis such as pCO2 monitoring system and environment monitoring systems.

  3. Combining transcutaneous blood gas measurement and pulse oximetry.

    Science.gov (United States)

    Eberhard, Patrick; Gisiger, P A; Gardaz, J P; Spahn, D R

    2002-01-01

    We are describing the preliminary results of tests performed in adult volunteers and in adult patients during and after general anesthesia with a miniaturized single sensor combining the continuous and non-invasive measurement of oxygen saturaiton by pulse oximetry (SpO2) and transcutaneous PCO2 (OxiCarbo sensor). The sensor is heated to 42 degrees C to arterialize the cutaneous tissue and is applied at the ear lobe with a special low-pressure clip. The results indicate a good agreement between ear lobe PCO2 and arterial PCO2 in the range 35 to 70 mmHg (10 patients, number of measurements 104, regression line TcPCO2 = 1.01 PaCO2 + 0.59 mmHg, bias 1.22 mmHg, SD 3.69 mmHg) and between ear lobe SpO2 and SaO2 (bias 0.44% with SD 0.77% in the range 80% to 100%, bias 1.39% with SD 1.43% in the range 60% to 80%). The ear lobe OxiCarbog sensor detects the SpO2 change 5 to 37 sec faster than a finger sensor and the PCO2 change 9 to 48 sec faster than a transcutaneous sensor fixed at the upper arm. Further improvements versus single sensors are a higher stability of the SpO2 signal and the possibility of performing long term SpO2 and PCO2 measurement at the ear lobe. PMID:11900043

  4. Validation of a Transcutaneous CO2 Monitor in Adult Patients with Chronic Respiratory Failure

    NARCIS (Netherlands)

    Hazenberg, A.; Zijlstra, J. G.; Kerstjens, H. A. M.; Wijkstra, P. J.

    2011-01-01

    Background: Home mechanical ventilation is usually started in hospital as arterial blood gas sampling is deemed necessary to monitor CO2 and O-2 adequately during institution of ventilatory support. A non-invasive device to reliably measure CO2 transcutaneously would alleviate the need for high care

  5. The accuracy of non-invasive carbon dioxide monitoring: a clinical evaluation of two transcutaneous systems.

    Science.gov (United States)

    Bolliger, D; Steiner, L A; Kasper, J; Aziz, O A; Filipovic, M; Seeberger, M D

    2007-04-01

    We determined the accuracy of two transcutaneous carbon dioxide monitoring systems (SenTec Digital Monitor with V-Sign Sensor and TOSCA 500 with TOSCA Sensor 92) for the measurement of single values and trends in the arterial partial pressure of carbon dioxide in 122 adult patients during major surgery and in 50 adult patients in the intensive care unit. One or several paired measurements were performed in each patient. The first measurement was used to determine the accuracy of a single value of transcutaneous carbon dioxide; the difference between the first and the last measurements was used to analyse the accuracy and to track trends. We defined a 95% limit of agreement of agreement between transcutaneous carbon dioxide partial pressure values derived from the two systems and arterial carbon dioxide values for both single values and trends as defined by our suggested limit of agreement. We conclude that these systems cannot replace conventional blood gas analysis in the clinical setting studied. PMID:17381578

  6. Non Invasive Transcutaneous Carbondioxide Monitoring in Adult Open Heart Surgery

    OpenAIRE

    UÇAR, P.; GAZİOĞLU, G.; ERDEMLİ, Ö.; ÇİÇEK, Ö. F.; DEMİR, A.

    2013-01-01

    Non Invasive Transcutaneous Carbondioxide Monitoring in Adult Open Heart SurgeryObjective: Follow-up of CO level during open heart surgery is crucial in terms of monitoring and management of the metabolic status. In this prospective observational study,measurement of end tidal CO and arterial CO levels in adult open heart surgery was compared with transcutaneous CO monitoring which is a non-invasive method.Material and Method: The study included 22 ASA II-III patients with an age range of 30-...

  7. Continuous intra-arterial blood gas monitoring. A clinical experience.

    Science.gov (United States)

    Paolillo, G; Tosoni, A; Mariani, M A; Venturino, M

    1994-01-01

    Miniaturized sensors, based upon the principles of optical fluorescence, can measure in vivo the pH, pCO2 value and pO2 value of blood. In this report we studied continuous intra-arterial blood gas monitoring in 27 patients undergoing cardiac surgery (no. 16 coronary artery by-pass grafting, no. 2 valvular surgery) and major vascular surgery (no. 9 abdominal aortic aneurysms). Total duration of continuous intra-arterial blood gas monitoring was 677 hours, with a ratio of 25.0 +/- 14.8 hours/patient (range 4-96 hours). The in vitro values of pH, pCO2 and pO2 were compared to simultaneous records from the fiberoptic sensor for each of the 283 arterial blood gas samples obtained, by means of linear regression and Bland-Altman method, in order to test the correlation and the agreement between the two methods of measuring. For pH average bias was -0.023 and intersensor precision was 0.028, with a strong correlation (R = 0.92; p sensor precision was 2.65, with a slight decrease in correlation (R = 0.89; p < 0.001) and agreement. For pO2 average bias was -2.69 and the intersensor precision was 12.16, with a strong correlation (R = 0.97; p < 0.001) and agreement. In addition, we tested the reliability of the system for values of pO2 above 100 mmHg and we found a strong correlation (R = 0.96; p20.001) and agreement even for these clinical conditions, largely out of physiologic parameters. This study demonstrates the feasibility and reliability of continuous intra-arterial three-component PB 3300 (Puritan Bennett) blood gas monitoring. PMID:7800182

  8. The influence of transcutaneous electrical neurostimulation (TENS) on human cerebral blood flow velocities

    NARCIS (Netherlands)

    ter Laan, Mark; van Dijk, J. Marc C.; Elting, Jan-Willem J.; Fidler, Vaclav; Staal, Michiel J.

    2010-01-01

    It has been shown that transcutaneous electrical neurostimulation (TENS) reduces sympathetic tone. Spinal cord stimulation (SCS) has proven qualities to improve coronary, peripheral, and cerebral blood circulation. Therefore, we postulate that TENS and SCS affect the autonomic nervous system in anal

  9. Carbon dioxide monitoring during laparoscopic-assisted bariatric surgery in severely obese patients: transcutaneous versus end-tidal techniques.

    Science.gov (United States)

    Dion, Joanna M; McKee, Chris; Tobias, Joseph D; Herz, Daniel; Sohner, Paul; Teich, Steven; Michalsky, Marc

    2015-02-01

    Various factors including severe obesity or increases in intra-abdominal pressure during laparoscopy can lead to inaccuracies in end-tidal carbon dioxide (PETCO2) monitoring. The current study prospectively compares ET and transcutaneous (TC) CO2 monitoring in severely obese adolescents and young adults during laparoscopic-assisted bariatric surgery. Carbon dioxide was measured with both ET and TC devices during insufflation and laparoscopic bariatric surgery. The differences between each measure (PETCO2 and TC-CO2) and the PaCO2 were compared using a non-paired t test, Fisher's exact test, and a Bland-Altman analysis. The study cohort included 25 adolescents with a mean body mass index of 50.2 kg/m2 undergoing laparoscopic bariatric surgery. There was no difference in the absolute difference between the TC-CO2 and PaCO2 (3.2±3.0 mmHg) and the absolute difference between the PETCO2 and PaCO2 (3.7±2.5 mmHg). The bias and precision were 0.3 and 4.3 mmHg for TC monitoring versus PaCO2 and 3.2 and 3.2 mmHg for ET monitoring versus PaCO2. In the young severely obese population both TC and PETCO2 monitoring can be used to effectively estimate PaCO2. The correlation of PaCO2 to TC-CO2 is good, and similar to the correlation of PaCO2 to PETCO2. In this population, both of these non-invasive measures of PaCO2 can be used to monitor ventilation and minimize arterial blood gas sampling.

  10. Application of transcutaneous diffuse reflectance spectroscopy in the measurement of blood glucose concentration

    Institute of Scientific and Technical Information of China (English)

    Wenliang Chen(陈文亮); Rong Liu(刘蓉); Houxin Cui(崔厚欣); Kexin Xu(徐可欣); Lina Lü(吕丽娜)

    2004-01-01

    In this paper, the propagation characteristics of near-infrared (NIR) light in the palm tissue are analyzed,and the principle and feasibility of using transcutaneous diffuse reflectance spectroscopy for non-invasive blood glucose detection are presented. An optical probe suitable for measuring the diffuse reflectance spectrum of human palm and a non-invasive blood glucose detection system using NIR spectroscopy are designed. Based on this system, oral glucose tolerance tests are performed to measure the blood glucose concentrations of two young healthy volunteers. The partial least square calibration model is then constructed by all individual experimental data. The final result shows that correlation coefficients of the two experiments between the predicted blood glucose concentrations and the reference blood glucose concentrations are 0.9870 and 0.9854, respectively. The root mean square errors of prediction of full cross validation are 0.54 and 0.52 mmol/1, respectively.

  11. Transcutaneous oximetry monitoring in patients with type 2 diabetes mellitus and critical limb ischemia

    Directory of Open Access Journals (Sweden)

    O N Bondarenko

    2013-03-01

    Full Text Available Aims. To evaluate transcutaneous oximetry as a method for diagnostics and monitoring in patients with diabetes mellitus (DM and critical limb ischemia (CLI after percutaneous transluminal balloon angioplasty (PTBA.Materials and Methods. We enrolled 126 patients with DM and CLI for participation in this study (148 limbs in total. 22 patients underwent PTBA on both lower limbs, and 104 – on single limb. Transcutaneous oximetry and duplex ultrasonography of lower limb arteries was performed prior to PTBA with subsequent examinations on 5-7th days, 1st, 3rd and 6th month after intervention. Transcu- taneous oxygen tension (TcpO2 was measured by Radiometer (Copenhagen oximeter system. Duplex ultrasonography was performed on Voluson 730® Expert system (GE Medical Systems Kretztechnik GmbH&Co OHG, Austria.Results. Multiple factor analysis suggests that results of TcpO2 monitoring prior to and after PTBA are influenced by presence of ischemic heart disease, severe lower limb infections, serum creatinine, arterial hypertension and lower limb reperfusion edema. We observed a strong correlation of TcpO2 with the degree of anterior tibial artery and dorsal pedis artery occlusion.Conclusion. Transcutaneous oximetry allows evaluation of CLI severity and efficiency of PTBA in the majority oа patients with DM and CLI. Certain comorbidities impose limitations on this technique. Efficiency of endovascular intervention should be evaluated based on complex non-invasive examination, clinical data and signs of CLI.

  12. Assessment of a continuous blood gas monitoring system in animals during circulatory stress

    Directory of Open Access Journals (Sweden)

    Renzulli Attilio

    2011-01-01

    Full Text Available Abstract Background The study was aimed to determine the measurement accuracy of The CDI™ blood parameter monitoring system 500 (Terumo Cardiovascular Systems Corporation, Ann Arbor MI in the real-time continuous measurement of arterial blood gases under different cardiocirculatory stress conditions Methods Inotropic stimulation (Dobutamine 2.5 and 5 μg/kg/min, vasoconstriction (Arginine-vasopressin 4, 8 and 16 IU/h, hemorrhage (-10%, -20%, -35%, and -50% of the theoretical volemia, and volume resuscitation were induced in ten swine (57.4 ± 10.7 Kg.Intermittent blood gas assessments were carried out using a routine gas analyzer at any experimental phase and compared with values obtained at the same time settings during continuous monitoring with CDI™ 500 system. The Bland-Altman analysis was employed. Results Bias and precision for pO2 were - 0.06 kPa and 0.22 kPa, respectively (r2 = 0.96; pCO2 - 0.02 kPa and 0.15 kPa, respectively; pH -0.001 and 0.01 units, respectively ( r2 = 0.96. The analysis showed very good agreement for SO2 (bias 0.04,precision 0.33, r2 = 0.95, Base excess (bias 0.04,precision 0.28, r2 = 0.98, HCO3 (bias 0.05,precision 0.62, r2 = 0.92,hemoglobin (bias 0.02,precision 0.23, r2 = 0.96 and K+ (bias 0.02, precision 0.27, r2 = 0.93. The sensor was reliable throughout the experiment during hemodynamic variations. Conclusions Continuous blood gas analysis with the CDI™ 500 system was reliable and it might represent a new useful tool to accurately and timely monitor gas exchange in critically ill patients. Nonetheless, our findings need to be confirmed by larger studies to prove its reliability in the clinical setting.

  13. Modulation of cerebral blood flow with transcutaneous electrical neurostimulation (TENS) in patients with cerebral vasospasm after subarachnoid hemorrhage

    NARCIS (Netherlands)

    ter Laan, Mark; van Dijk, J.M.C.; Stewart, Roy; Staal, Michiel J; Elting, Jan-Willem J

    2014-01-01

    ObjectivesTranscutaneous electrical neurostimulation (TENS) and spinal cord stimulation have been shown to increase peripheral and cerebral blood flow. We postulate that certain pathological conditions attenuate cerebral autoregulation, which may result in a relative increase of the importance of ne

  14. Non-arterial assessment of blood gas status in patients with chronic pulmonary disease.

    OpenAIRE

    Elborn, J. S.; Finch, M B; Stanford, C. F.

    1991-01-01

    Assessment of blood gas status is important in the management of patients with chronic pulmonary disease. Arterial puncture is often painful and may damage the arterial wall. Measurement of oxygen saturation by transcutaneous oximetry offers a non-invasive alternative to arterial methods but does not allow assessment of partial pressure of carbon dioxide. We have examined the value of oximetry and dorsal hand venous carbon dioxide as an alternative to arterial puncture. Transcutaneous oxygen ...

  15. Continuous arterial blood gas monitoring in rabbits: an efficient method for evaluation of ratio-based optrodes

    Science.gov (United States)

    Martin, Roy C.; Olstein, Alan D.; Malin, Stephen F.; Perkovich, Anne

    1992-04-01

    Laboratory bench testing of optical blood gas sensors is insufficient to completely predict capabilities. Sensor testing in animals offers advantages of known physiologic and regulatory mechanisms of hemodynamics to better predict sensor performance. The domestic rabbit, Oryctalogis Cuniculus, a lagomorph of the family Leporidae was used for sensor evaluation. The rabbits are ventilated and blood gases modulated by variations in FIO2 and rate adjustments. Twenty gauge catheters are placed in the dorsal aorta, cartoid, and femoral arteries. Pressures are monitored via transducers on the arterial lines. The optical blood gas sensors are fitted within the catheters and blood samples are collected over them for bench analysis. Sensors are on 125 micrometers glass optic fibers. Proprietary prepolymers are applied on the fiber tips through in fiber photopolymerization. These sensors are then calibrated in tonometered water and blood. Sensor monitoring is accomplished through OSR microfluorimetry systems. We have used this model in 26 studies over the past six months evaluating over fifty blood gas sensors. These studies have lasted from six to twenty-four hours. Our correlation of sensor readings to assayed blood samples is r2 equals .97 for pH values of 6.80 - 7.70, r2 equals .94 for PCO2 values of 10 - 175 mmHg and r2 equals .94 for PO2 values of 10 - 350 mmHg.

  16. Monitor blood glucose - slideshow

    Science.gov (United States)

    ... medlineplus.gov/ency/presentations/100220.htm Monitoring blood glucose - Series—Monitoring blood glucose: Using a self-test meter To use the ... A.M. Editorial team. Related MedlinePlus Health Topics Blood Sugar A.D.A.M., Inc. is accredited by ...

  17. Transcutaneous cervical esophagus ultrasound in adults: Relation with ambulatory 24-h pH-monitoring and esophageal manometry

    OpenAIRE

    Kacar, Sabite; Uysal, Selma; Kuran, Sedef; Dagli, Ulku; Ozin, Yasemin; Karabulut, Erdem; Sasmaz, Nurgul

    2007-01-01

    AIM: To determine the gastroesophageal refluxate in the cervical esophagus (CE) and measure transcutaneous cervical esophageal ultrasound (TCEUS) findings [anterior wall thickness (WT) of CE, esophageal luminal diameter (ELD), esophageal diameter (ED)]; to compare TCEUS findings in the patient subgroups divided according to 24-h esophageal pH monitoring and manometry; and to investigate possible cut-off values according to the TCEUS findings as a predictor of gastroesophageal reflux (GER).

  18. Vasogen's immune modulation therapy (IMT) improves postischemic foot skin blood flow and transcutaneous pO(2) recovery rates in patients with advanced peripheral arterial occlusive disease

    DEFF Research Database (Denmark)

    Edvinsson, L I; Edvinsson, M L; Angus Deveber, G

    2003-01-01

    or Vasogen's IMT over a 9-week period. Dorsal foot skin blood flow was assessed directly using laser Doppler fluxmetry (LDF) and indirectly using measurement of transcutaneous pO(2) (tcpO(2)). Key outcome measures of skin blood flow were, for LDF: resting values, peak postischemic values, and the total time...

  19. Home monitoring of blood pressure

    OpenAIRE

    McGrath, Barry P

    2015-01-01

    Home blood pressure monitoring is the self-measurement of blood pressure by patients. In the diagnosis and management of high blood pressure it is complementary to 24-hour ambulatory blood pressure monitoring and clinic blood pressure measurements. Home monitoring can also help to identify white-coat and masked hypertension.

  20. [Changes in the blood rheological properties in the transcutaneous irradiation of the ulnar vascular fascicle with a helium-neon laser].

    Science.gov (United States)

    Paleev, N R; Karandashov, V I; Voronina, M A; Fin'ko, I A

    1993-10-01

    An investigation of blood rheologic properties in 12 patients with acute pneumonia has been made by using low-energy He-Ne laser (LG-79-2) irradiation of vascular fascicle. 37 exposures have been performed, 40 min each. Immediate effects of He-Ne laser were studied comparing blood samples taken before and after the radiation. Results obtained have demonstrated transcutaneous blood irradiation causing prompt and pronounced effects on blood rheologic characteristics: reduced blood viscosity, improved both viscous-elastic properties and osmotic resistance of erythrocytes, activated platelet aggregation. These effects excluding the latter are rather positive for the human body. As for platelet aggregation, its activation following transcutaneous laser irradiation of blood might be expected to grow into a factor of risk provoking pathologic thrombogenesis in venous congestion, hypercoagulation and vascular wall injury.

  1. Effects of the transcutaneous electrode temperature on the accuracy of transcutaneous carbon dioxide tension

    DEFF Research Database (Denmark)

    Sørensen, Line C; Brage-Andersen, Lene; Greisen, Gorm

    2011-01-01

    The harmful effect of hypocapnia on the neonatal brain emphasizes the importance of monitoring arterial carbon dioxide tension (PaCO2). Transcutaneous monitoring of carbon dioxide (tcPCO2) reduces the need for arterial blood sampling. Drawbacks are high electrode temperature causing risks of skin...... burning. The aim was to determine the accuracy and precision of tcPCO2 at reduced electrode temperature....

  2. Real-time gastric motility monitoring using transcutaneous intraluminal impedance measurements (TIIM)

    International Nuclear Information System (INIS)

    The stomach plays a critical role in digestion, processing ingested food mechanically and breaking it up into particles, which can be effectively and efficiently processed by the intestines. When the motility of the stomach is compromised, digestion is adversely affected. This can lead to a variety of disorders. Current diagnostic techniques for gastric motility disorders are seriously lacking, and are based more on eliminating other possibilities rather than on specific tests. Presently, gastric motility can be assessed by monitoring gastric emptying, food transit, intragastric pressures, etc. The associated tests are usually stationary and of relatively short duration. The present study proposes a new method of measuring gastric motility, utilizing the attenuation of an oscillator-induced electrical signal across the gastric tissue, which is modulated by gastric contractions. The induced high-frequency oscillator signal is generated within the stomach, and is picked up transluminally by cutaneous electrodes positioned on the abdominal area connected to a custom-designed data acquisition instrument. The proposed method was implemented in two different designs: first a transoral catheter was modified to emit the signal inside the stomach; and second, a gastric retentive pill was designed to emit the signal. Both implementations were applied in vivo on two mongrel dogs (25.50 kg and 25.75 kg). Gastric contractions were registered and quantitatively compared to recordings from force transducers sutured onto the serosa of the stomach. Gastric motility indices were calculated for each minute, with transluminal impedance measurements and the measurements from the force transducers showing statistically significant (p < 0.05) Pearson correlation coefficients (0.65 ± 0.08 for the catheter-based design and 0.77 ± 0.03 for the gastric retentive pill design). These results show that transcutaneous intraluminal impedance measurement has the potential with further research

  3. Blood gas analysis for bedside diagnosis

    OpenAIRE

    Singh, Virendra; Khatana, Shruti; Gupta, Pranav

    2013-01-01

    Arterial blood gas is an important routine investigation to monitor the acid-base balance of patients, effectiveness of gas exchange, and the state of their voluntary respiratory control. Majority of the oral and maxillofacial surgeons find it difficult to interpret and clinically correlate the arterial blood gas report in their everyday practice. This has led to underutilization of this simple tool. The present article aims to simplify arterial blood gas analysis for a rapid and easy bedside...

  4. Gas exchange as monitored in mixed venous and arterial blood during experimental cardiopulmonary resuscitation.

    Science.gov (United States)

    Wiklund, L; Jorfeldt, L; Stjernström, H; Rubertsson, S

    1992-07-01

    Nineteen anaesthetized piglets were investigated. After catheterization and a stabilization period, ventricular fibrillation was induced with a transthoracic DC shock, after which a 10-min period of cardiopulmonary resuscitation (CPR) took place. CPR included manual chest compression and mechanical ventilation with pure oxygen. After 1 min of CPR, an infusion of alkaline buffer was begun and completed within 5 min. A total of 50 mmol of either sodium bicarbonate (n = 6) or tris buffer mixture (n = 7) were given. These two groups were compared with a third control group (n = 6) receiving the same volume of normal saline. After 8 min of CPR all animals were given 0.5 mg adrenaline i.v., and after 10 min DC shocks were used to revert the heart back to normal sinus rhythm. Our results demonstrate that blood flow and not ventilation is the limiting factor for the efficient disposal of CO2 during CPR. This also applied when the demand for CO2 transport was increased by administration of sodium bicarbonate. The respiratory exchange ratio increased 1.9-fold, indicating that the transport of carbon dioxide was less affected than that of oxygen. The estimated alveolo-arterial oxygen tension difference, shunt, and overall ventilation/perfusion ratio increased, creating an inverse hyperbolic relationship between arterial PCO2 and PO2. The difference between mixed venous and arterial PCO2 correlated well to the mixed venous PCO2, implying more efficient pulmonary elimination of PCO2 when the mixed venous PCO2 was high. Pulmonary gas exchange during CPR appears to be independent of alkaline buffer therapy in the form of sodium bicarbonate or tris buffer mixture.(ABSTRACT TRUNCATED AT 250 WORDS) PMID:1632165

  5. S35: a new parameter in blood gas analysis for monitoring the systemic oxygenation.

    Science.gov (United States)

    Trouwborst, A; Tenbrinck, R; Van Woerkens, E C

    1990-01-01

    In the estimation of oxygen transport the term oxygen availability is used as the product of cardiac output and the arterial oxygen content (CaO2). Attempts can be made to modify the concept of oxygen availability by subtracting from the CaO2 the venous content at a critical PO2 as measured in mixed venous blood (Pv-O2), where oxygen diffusion into tissue becomes compromised and oxygen uptake (VO2) may decrease. The real arterial available oxygen content (CavlO2) can be calculated by estimating the saturation at the critical Pv-O2. For our concept S35 was chosen as such a dynamic baseline. Similar modification of oxygen extraction ratio (ERav) defined as VO2 divided by the real oxygen availability (O2av) should give, more than the classic ER, a realistic indices of oxygen availability in relation to oxygen consumption. It can be hypothesized that VO2 starts to decline when ERav is around 1.0. During isovolemic hemodilution VO2 started to drop when ERav reached 1.08 +/- 0.09. The S35 changed from 55.0 +/- 2.1% to 41.5 +/- 4.1%, correlated with changes in Pv-O2. A direct correlation was also found between the increase of the classic ER and the change in S35. We conclude that the S35, the CavlO2 and the ERav can be of value in monitoring the systemic oxygenation and that the concept also includes the effect of changes in oxyhemoglobin characteristics on oxygen delivery.

  6. Noninvasive Transcutaneous Monitoring in Long-Term Follow-Up of Patients With Thromboangiitis Obliterans Treated With Intravenous Iloprost.

    Science.gov (United States)

    Melillo, Elio; Grigoratos, Chrysanthos; Sanctis, Francesco De; Spontoni, Paolo; Nuti, Marco; Dell'Omodarme, Matteo; Ferrari, Mauro; Balbarini, Alberto

    2015-07-01

    We evaluated the effectiveness of intravenous iloprost (IVI) in outpatients with thromboangiitis obliterans (TAO) and lower limb noninvasive transcutaneous monitoring (TCM) at follow-up (FU). Ten consecutive patients with TAO underwent IVI therapy. Transcutaneous oxygen (TcPo 2) and carbon dioxide (TcPco 2) determination and laser Doppler flowmetry (LDF) were performed before and after IVI at 3, 6, and 12 months of FU. Clinical response was positive in 7 patients, whereas 3 nonresponders underwent a second IVI cycle with 1 showing a late positive clinical response. After 12 months of FU, all patients were alive without amputations. Supine and dependent TcP2 levels significantly improved (P < .005). Hallux LDF values showed significant change with the maximal hyperemic test at 44°C (P < .005). Forefoot maximal hyperemic test at 44°C LDF (P < .005) and improved venous arterial reflex (P < .05) showed statistically significant time evolution. We demonstrated some degree of IVI effectiveness and evaluated TCM in patients with TAO.

  7. DIGITAL BLOOD PRESSURE MONITOR

    Directory of Open Access Journals (Sweden)

    R. Fuentes

    2004-12-01

    Full Text Available In this work we present a blood pressure monitor which measures both the high blood pressure (systolic pressure,and the low blood pressure (diastolic pressure. It is a semiautomatic meter because the inflation of the occlusivecuff is carried out in a manual way. The transducer used is a piezoresistive silicon pressure sensor integrated onchip which provides a proportional voltage to the input pressure, with a measurement range from 0 to 50 kPa (0–7.3 PSI. The oscillometric method is employed, which consists on detecting the oscillometric signal on brachialartery, being processed at each pressure step, when the cuff is gradually deflated. Signal sampling is carried out ata rate determined by the heart rate.In order to program the digital electronics of the circuit we used Altera tools, with the compiler MAX-PLUS II, andthe device selected to implement the design was an EPM7128SLC84-15 CPLD (Complex Programmable LogicDevice

  8. A novel system for transcutaneous application of carbon dioxide causing an "artificial Bohr effect" in the human body.

    Directory of Open Access Journals (Sweden)

    Yoshitada Sakai

    Full Text Available BACKGROUND: Carbon dioxide (CO(2 therapy refers to the transcutaneous administration of CO(2 for therapeutic purposes. This effect has been explained by an increase in the pressure of O(2 in tissues known as the Bohr effect. However, there have been no reports investigating the oxygen dissociation of haemoglobin (Hb during transcutaneous application of CO(2in vivo. In this study, we investigate whether the Bohr effect is caused by transcutaneous application of CO2 in human living body. METHODS: We used a novel system for transcutaneous application of CO(2 using pure CO(2 gas, hydrogel, and a plastic adaptor. The validity of the CO(2 hydrogel was confirmed in vitro using a measuring device for transcutaneous CO(2 absorption using rat skin. Next, we measured the pH change in the human triceps surae muscle during transcutaneous application of CO(2 using phosphorus-31 magnetic resonance spectroscopy ((31P-MRS in vivo. In addition, oxy- and deoxy-Hb concentrations were measured with near-infrared spectroscopy in the human arm with occulted blood flow to investigate O2 dissociation from Hb caused by transcutaneous application of CO(2. RESULTS: The rat skin experiment showed that CO(2 hydrogel enhanced CO(2 gas permeation through the rat skin. The intracellular pH of the triceps surae muscle decreased significantly 10 min. after transcutaneous application of CO(2. The NIRS data show the oxy-Hb concentration decreased significantly 4 min. after CO(2 application, and deoxy-Hb concentration increased significantly 2 min. after CO(2 application in the CO(2-applied group compared to the control group. Oxy-Hb concentration significantly decreased while deoxy-Hb concentration significantly increased after transcutaneous CO(2 application. CONCLUSIONS: Our novel transcutaneous CO(2 application facilitated an O(2 dissociation from Hb in the human body, thus providing evidence of the Bohr effect in vivo.

  9. Electrical characteristic of the titanium mesh electrode for transcutaneous intrabody communication to monitor implantable artificial organs.

    Science.gov (United States)

    Okamoto, Eiji; Kikuchi, Sakiko; Mitamura, Yoshinori

    2016-09-01

    We have developed a tissue-inducing electrode using titanium mesh to obtain mechanically and electrically stable contact with the tissue for a new transcutaneous communication system using the human body as a conductive medium. In this study, we investigated the electrical properties of the titanium mesh electrode by measuring electrode-tissue interface resistance in vivo. The titanium mesh electrode (Hi-Lex Co., Zellez, Hyogo, Japan) consisted of titanium fibers (diameter of 50 μm), and it has an average pore size of 200 μm and 87 % porosity. The titanium mesh electrode has a diameter of 5 mm and thickness of 1.5 mm. Three titanium mesh electrodes were implanted separately into the dorsal region of the rat. We measured the electrode-electrode impedance using an LCR meter for 12 weeks, and we calculated the tissue resistivity and electrode-tissue interface resistance. The electrode-tissue interface resistance of the titanium mesh electrode decreased slightly until the third POD and then continuously increased to 75 Ω. The electrode-tissue interface resistance of the titanium mesh electrode is stable and it has lower electrode-tissue interface resistance than that of a titanium disk electrode. The extracted titanium mesh electrode after 12 weeks implantation was fixed in 10 % buffered formalin solution and stained with hematoxylin-eosin. Light microscopic observation showed that the titanium mesh electrode was filled with connective tissue, inflammatory cells and fibroblasts with some capillaries in the pores of the titanium mesh. The results indicate that the titanium mesh electrode is a promising electrode for the new transcutaneous communication system.

  10. Variability of Transcutaneous Oxygen and Carbon Dioxide Pressure Measurements Associated with Sensor Location.

    Science.gov (United States)

    Górska, K; Korczyński, P; Maskey-Warzęchowska, M; Chazan, R; Krenke, R

    2015-01-01

    Transcutaneous measurement of oxygen and carbon dioxide pressure (PtcO2 and PtcCO2) is useful in gas exchange monitoring. However, the relationship between PtcO2, pulse oximetry (SaO2) and arterial blood gases (ABG) is unclear. The aim of the present study was to compare PtcO2 and PtcCO2 with SaO2 and ABG, to evaluate the effect of sensor location on the results and stability of PtcO2 and PtcCO2, and to assess the impact of body composition on PtcO2 and PtcCO2. PtcO2 and PtcCO2 were measured in 20 healthy volunteers at three locations: right second intercostal space, lateral surface of the abdomen, and the inner surface of the left arm. The results were recorded 10, 15, and 20 min after sensor fixation and compared with SaO2 and ABG measured 20 min after electrode placement on the chest. Body composition was evaluated by bioimpedance. The findings were that PtcO2 was stable on the chest; but on the arm and abdomen it increased and reached maximum at 20 min. Transcutaneous PCO2 stabilized at 10 min in all the three locations. No significant correlations between PtcO2 and SaO2 or PaO2 were found. Transcutaneous PCO2 correlated with PaCO2. Both PtcO2 and PtcCO2 were not influenced by body composition. We conclude that the value of PtcO2 in monitoring of blood oxygenation was not unequivocally confirmed; PtcCO2 reliably reflects PaCO2, irrespective of sensor location. Body composition does not affect PtcO2 and PtcCO2. PMID:25820668

  11. Monitoring Blood Sugar: The Importance of Checking Blood Sugar Levels

    Science.gov (United States)

    ... 5 Things to Know About Zika & Pregnancy Monitoring Blood Sugar KidsHealth > For Parents > Monitoring Blood Sugar Print ... Other Tests Record Keeping The Importance of Checking Blood Sugar Levels Besides helping to keep blood sugar ...

  12. Transcutaneous Intraluminal Impedance Measurement for Minimally Invasive Monitoring of Gastric Motility: Validation in Acute Canine Models

    Directory of Open Access Journals (Sweden)

    Michael D. Poscente

    2014-01-01

    Full Text Available Transcutaneous intraluminal impedance measurement (TIIM is a new method to cutaneously measure gastric contractions by assessing the attenuation dynamics of a small oscillating voltage emitted by a battery-powered ingestible capsule retained in the stomach. In the present study, we investigated whether TIIM can reliably assess gastric motility in acute canine models. Methods. Eight mongrel dogs were randomly divided into 2 groups: half received an active TIIM pill and half received an identically sized sham capsule. After 24-hour fasting and transoral administration of the pill (active or sham, two force transducers (FT were sutured onto the antral serosa at laparotomy. After closure, three standard cutaneous electrodes were placed on the abdomen, registering the transluminally emitted voltage. Thirty-minute baseline recordings were followed by pharmacological induction of gastric contractions using neostigmine IV and another 30-minute recording. Normalized one-minute baseline and post-neostigmine gastric motility indices (GMIs were calculated and Pearson correlation coefficients (PCCs between cutaneous and FT GMIs were obtained. Statistically significant GMI PCCs were seen in both baseline and post-neostigmine states. There were no significant GMI PCCs in the sham capsule test. Further chronic animal studies of this novel long-term gastric motility measurement technique are needed before testing it on humans.

  13. The application of transcutaneous CO2 pressure monitoring in the anesthesia of obese patients undergoing laparoscopic bariatric surgery.

    Directory of Open Access Journals (Sweden)

    Shijiang Liu

    Full Text Available To investigate the correlation and accuracy of transcutaneous carbon dioxide partial pressure (PTCCO2 with regard to arterial carbon dioxide partial pressure (PaCO2 in severe obese patients undergoing laparoscopic bariatric surgery. Twenty-one patients with BMI>35 kg/m(2 were enrolled in our study. Their PaCO2, end-tidal carbon dioxide partial pressure (PetCO2, as well as PTCCO2 values were measured at before pneumoperitoneum and 30 min, 60 min, 120 min after pneumoperitoneum respectively. Then the differences between each pair of values (PetCO2-PaCO2 and. (PTCCO2-PaCO2 were calculated. Bland-Altman method, correlation and regression analysis, as well as exact probability method and two way contingency table were employed for the data analysis. 21 adults (aged 19-54 yr, mean 29, SD 9 yr; weight 86-160 kg, mean 119.3, SD 22.1 kg; BMI 35.3-51.1 kg/m(2, mean 42.1,SD 5.4 kg/m(2 were finally included in this study. One patient was eliminated due to the use of vaso-excitor material phenylephrine during anesthesia induction. Eighty-four sample sets were obtained. The average PaCO2-PTCCO2 difference was 0.9 ± 1.3 mmHg (mean ± SD. And the average PaCO2-PetCO2 difference was 10.3 ± 2.3 mmHg (mean ± SD. The linear regression equation of PaCO2-PetCO2 is PetCO2 = 11.58+0.57 × PaCO2 (r(2 = 0.64, P<0.01, whereas the one of PaCO2-PTCCO2 is PTCCO2 = 0.60 + 0.97 × PaCO2 (r(2 = 0.89. The LOA (limits of agreement of 95% average PaCO2-PetCO2 difference is 10.3 ± 4.6 mmHg (mean ± 1.96 SD, while the LOA of 95% average PaCO2-PTCCO2 difference is 0.9 ± 2.6 mmHg (mean ± 1.96 SD. In conclusion, transcutaneous carbon dioxide monitoring provides a better estimate of PaCO2 than PetCO2 in severe obese patients undergoing laparoscopic bariatric surgery.

  14. Transcutaneous oxygen tension in imminent foot gangrene

    DEFF Research Database (Denmark)

    Tønnesen, K H

    1978-01-01

    Transcutaneous oxygen tension at 44 degree C and maximal isotope clearance (90m Tc-pretechnetate + histramine) just proximal to the 1st toe and systolic toe blood pressure (strain gauge) were studied on a tilt table in patients with various degrees of obstructive arteriosclerotic disease. In legs...... with moderate obstruction, the oxygen tension reached zero at a toe systolic blood pressure of 5--10 mmHg (tilt toe up) and reached arterial oxygen tension at about 50 to 70 mmHg (tilt toe down). In legs withsevere arterial obstruction and ischaemic rest pain, oxygen tension rose from zero not before systolic...... toe blood pressure reached 20--50 mmHg. Significant isotope clearance was seen at pressures below the limits just mentioned for both types of patients. This phenomenon here seen of a perfusion without oxygen supply is explained by a gas leak (rendered significant because of the slow flow rate) from...

  15. Transcutaneous cervical esophagus ultrasound in adults: Relation with ambulatory 24-h pH-monitoring and esophageal manometry

    Institute of Scientific and Technical Information of China (English)

    2007-01-01

    AIM: To determine the gastroesophageal refluxate in the cervical esophagus (CE) and measure transcutaneous cervical esophageal ultrasound (TCEUS) findings [anterior wall thickness (WT) of CE, esophageal luminal diameter (ELD), esophageal diameter (ED)]; to compare TCEUS findings in the patient subgroups divided according to 24-h esophageal pH monitoring and manometry; and to investigate possible cut-off values according to the TCEUS findings as a predictor of gastroesophageal reflux (GER).METHODS: In 45/500 patients, refluxate was visualized in TCEUS. 38/45 patients underwent esophagogastroduo denoscopy (EGD), 24-h pH monitoring and manometry.RESULTS: The 38 patients were grouped according to 24-h pH monitoring as follows: Group A: GER-positive (n = 20) [Includes Group B: isolated proximal reflux (PR) (n = 6), Group C: isolated distal reflux (DR) (n = 6),and Group D: both PR/DR (n = 8)]; Group E: no reflux (n = 13); and Group F: hypersensitive esophagus (HSE) (n= 5). Groups B + D indicated total PR patients (n = 14),Groups E + F reflux-negatives with HSE (n = 18), and Groups A + F reflux-positives with HSE (n = 25). When the 38 patients were grouped according to manometry findings, 24 had normal esophageal manometry; 7 had hypotensive and 2 had hypertensive lower esophageal sphincter (LES); and 5 had ineffective esophageal motility disorder (IEM). The ELD measurement was greater in group A + F than group E (P = 0.023, 5.0 ± 1.3 vs 3.9 ± 1.4 mm). In 27/38 patients, there was at least one pathologic acid reflux and/or pathologic manometry finding. The cut-off value for ELD of 4.83 mm had 79% sensitivity and 61% specificity in predicting the PR between Groups B + D and E (AUC = 0.775, P = 0.015).CONCLUSION: Visualizing refluxate in TCEUS was useful as a pre-diagnostic tool for estimating GER or manometric pathology in 71.1% of adults in our study,but it was not diagnostic for CE WT.

  16. Noninvasive monitoring of PaCO2 during one-lung ventilation and minimal access surgery in adults: End-tidal versus transcutaneous techniques

    OpenAIRE

    Cox, Paul; Tobias, Joseph D.

    2007-01-01

    Background: Previous studies have suggested that end-tidal CO2 (ET-CO2) may be inaccurate during one-lung ventilation (OLV). This study was performed to compare the accuracy of the noninvasive monitoring of PCO2 using transcutaneous CO2 (TC-CO2) with ET-CO2 in patients undergoing video-assisted thoracoscopic surgery (VATS) during OLV. Materials and Methods: In adult patients undergoing thoracoscopic surgical procedures, PCO2 was simultaneously measured with TC-CO2 and ET-CO2 devices and compa...

  17. Rugged and breathable forms of stretchable electronics with adherent composite substrates for transcutaneous monitoring

    Science.gov (United States)

    Jang, Kyung-In; Han, Sang Youn; Xu, Sheng; Mathewson, Kyle E.; Zhang, Yihui; Jeong, Jae-Woong; Kim, Gwang-Tae; Webb, R. Chad; Lee, Jung Woo; Dawidczyk, Thomas J.; Kim, Rak Hwan; Song, Young Min; Yeo, Woon-Hong; Kim, Stanley; Cheng, Huanyu; Rhee, Sang Il; Chung, Jeahoon; Kim, Byunggik; Chung, Ha Uk; Lee, Dongjun; Yang, Yiyuan; Cho, Moongee; Gaspar, John G.; Carbonari, Ronald; Fabiani, Monica; Gratton, Gabriele; Huang, Yonggang; Rogers, John A.

    2014-09-01

    Research in stretchable electronics involves fundamental scientific topics relevant to applications with importance in human healthcare. Despite significant progress in active components, routes to mechanically robust construction are lacking. Here, we introduce materials and composite designs for thin, breathable, soft electronics that can adhere strongly to the skin, with the ability to be applied and removed hundreds of times without damaging the devices or the skin, even in regions with substantial topography and coverage of hair. The approach combines thin, ultralow modulus, cellular silicone materials with elastic, strain-limiting fabrics, to yield a compliant but rugged platform for stretchable electronics. Theoretical and experimental studies highlight the mechanics of adhesion and elastic deformation. Demonstrations include cutaneous optical, electrical and radio frequency sensors for measuring hydration state, electrophysiological activity, pulse and cerebral oximetry. Multipoint monitoring of a subject in an advanced driving simulator provides a practical example.

  18. Rugged and breathable forms of stretchable electronics with adherent composite substrates for transcutaneous monitoring.

    Science.gov (United States)

    Jang, Kyung-In; Han, Sang Youn; Xu, Sheng; Mathewson, Kyle E; Zhang, Yihui; Jeong, Jae-Woong; Kim, Gwang-Tae; Webb, R Chad; Lee, Jung Woo; Dawidczyk, Thomas J; Kim, Rak Hwan; Song, Young Min; Yeo, Woon-Hong; Kim, Stanley; Cheng, Huanyu; Rhee, Sang Il; Chung, Jeahoon; Kim, Byunggik; Chung, Ha Uk; Lee, Dongjun; Yang, Yiyuan; Cho, Moongee; Gaspar, John G; Carbonari, Ronald; Fabiani, Monica; Gratton, Gabriele; Huang, Yonggang; Rogers, John A

    2014-09-03

    Research in stretchable electronics involves fundamental scientific topics relevant to applications with importance in human healthcare. Despite significant progress in active components, routes to mechanically robust construction are lacking. Here, we introduce materials and composite designs for thin, breathable, soft electronics that can adhere strongly to the skin, with the ability to be applied and removed hundreds of times without damaging the devices or the skin, even in regions with substantial topography and coverage of hair. The approach combines thin, ultralow modulus, cellular silicone materials with elastic, strain-limiting fabrics, to yield a compliant but rugged platform for stretchable electronics. Theoretical and experimental studies highlight the mechanics of adhesion and elastic deformation. Demonstrations include cutaneous optical, electrical and radio frequency sensors for measuring hydration state, electrophysiological activity, pulse and cerebral oximetry. Multipoint monitoring of a subject in an advanced driving simulator provides a practical example.

  19. Real-time gastric motility monitoring using transcutaneous intraluminal impedance measurements (TIIM).

    Science.gov (United States)

    Poscente, M D; Wang, G; Filip, D; Ninova, P; Yadid-Pecht, O; Andrews, C N; Mintchev, M P

    2014-02-01

    The stomach plays a critical role in digestion, processing ingested food mechanically and breaking it up into particles, which can be effectively and efficiently processed by the intestines. When the motility of the stomach is compromised, digestion is adversely affected. This can lead to a variety of disorders. Current diagnostic techniques for gastric motility disorders are seriously lacking, and are based more on eliminating other possibilities rather than on specific tests. Presently, gastric motility can be assessed by monitoring gastric emptying, food transit, intragastric pressures, etc. The associated tests are usually stationary and of relatively short duration. The present study proposes a new method of measuring gastric motility, utilizing the attenuation of an oscillator-induced electrical signal across the gastric tissue, which is modulated by gastric contractions. The induced high-frequency oscillator signal is generated within the stomach, and is picked up transluminally by cutaneous electrodes positioned on the abdominal area connected to a custom-designed data acquisition instrument. The proposed method was implemented in two different designs: first a transoral catheter was modified to emit the signal inside the stomach; and second, a gastric retentive pill was designed to emit the signal. Both implementations were applied in vivo on two mongrel dogs (25.50 kg and 25.75 kg). Gastric contractions were registered and quantitatively compared to recordings from force transducers sutured onto the serosa of the stomach. Gastric motility indices were calculated for each minute, with transluminal impedance measurements and the measurements from the force transducers showing statistically significant (p impedance measurement has the potential with further research and development to become a useful diagnostic technique. PMID:24398539

  20. Accuracy of Transcutaneous Carbon Dioxide Measurement in Premature Infants.

    Science.gov (United States)

    Janaillac, Marie; Labarinas, Sonia; Pfister, Riccardo E; Karam, Oliver

    2016-01-01

    Background. In premature infants, maintaining blood partial pressure of carbon dioxide (pCO2) value within a narrow range is important to avoid cerebral lesions. The aim of this study was to assess the accuracy of a noninvasive transcutaneous method (TcpCO2), compared to blood partial pressure of carbon dioxide (pCO2). Methods. Retrospective observational study in a tertiary neonatal intensive care unit. We analyzed the correlation between blood pCO2 and transcutaneous values and the accuracy between the trends of blood pCO2 and TcpCO2 in all consecutive premature infants born at weight: 1250 g), providing 1365 pairs of TcpCO2 and blood pCO2 values. Pearson's R correlation between these values was 0.58. The mean bias was -0.93 kPa with a 95% confidence limit of agreement of -4.05 to +2.16 kPa. Correlation between the trends of TcpCO2 and blood pCO2 values was good in only 39.6%. Conclusions. In premature infants, TcpCO2 was poorly correlated to blood pCO2, with a wide limit of agreement. Furthermore, concordance between trends was equally low. We warn about clinical decision-making on TcpCO2 alone when used as continuous monitoring. PMID:27375901

  1. Noninvasive continuous blood pressure monitoring

    Science.gov (United States)

    Poghosyan, Armen; Mouradian, Vahram; Hovhannisyan, Levon

    2015-03-01

    We are presenting a novel photoplethysmographic (PPG) optical sensor and device with ambient optical, electrical and electromagnetic noises cancellation, thus allowing only the useful optical signals to be received by the health monitoring device. We are also presenting a new processing technique for canceling the ambient noises contributed by optical, electrical and electromagnetic artifacts in the measured PPG signals. Such a device and method allow the enhancement of the performance of the PPG sensors compared to conventional apparatus and methods. The presented sensor and methodology have been integrated into a prototype standalone device for noninvasive, continuous, wearable, remote and mobile monitoring of blood pressure and other human vital signs, such as heart rate, oxygen saturation, respiration rate, etc This small device allows the user to read, store, process and transmit all the measurements made using the PPG optical sensor and the electronic unit to a remote location.

  2. Design of a prospective clinical study on the agreement between the Continuous GlucoseMonitor, a novel device for CONTinuous ASSessment of blood GLUcose levels, and the RAPIDLab® 1265 blood gas analyser: The CONTASSGLU study

    Directory of Open Access Journals (Sweden)

    Zimmermann Johannes B

    2012-09-01

    Full Text Available Abstract Background Although a device is needed to continuously measure blood glucose levels within an intensive care setting, and several large-scale prospective studies have shown that patients might benefit from intensive insulin, potassium, or glucose therapy during intensive care, no devices are currently available to continuously assess blood glucose levels in critically ill patients. We conceived the study described here to evaluate the clinical use of the Continuous Glucose Monitor (CGM performed via a central vein, and to determine the impact of phenomena, such as drift and shift, on the agreement between the CGM and a RAPIDLab® 1265 blood gas analyser (BGA. Methods/design In the CONTinuous ASSessment of blood GLUcose (CONTASSGLU study, up to 130 patients under intensive care will be fitted with the CGM, an ex vivo device that continuously measures blood glucose and lactate levels. Readings from the device taken 8 h after initial placement and calibration will be compared with values measured by a BGA. For this study, we chose the BGA as it is an established standard point-of-care device, instead of the devices used in certified central laboratories. Nevertheless, we will also independently compare the results from the point-of-care BGA with those determined by a central laboratory-based device. Blood samples will be collected from each patient from the same site in which the CGM will measure blood glucose. Consequently, each participant will serve as their own control, and no randomisation is necessary. The 95% limits of agreement and the corresponding confidence intervals will be calculated and compared with a prespecified clinically acceptable relative difference of 20%. Discussion Several attempts have been made to develop a device to continuously measure blood glucose levels within an intensive care setting or to use the devices that were originally designed for diabetes management, as several of these devices are already

  3. Home blood pressure monitoring for mild hypertensives.

    OpenAIRE

    Midanik, L T; Resnick, B; Hurley, L B; Smith, E J; Mccarthy, M.

    1991-01-01

    A clinical trial of 204 untreated patients with mild hypertension was conducted to assess the effect of home blood pressure monitoring on blood pressure level, pharmacologic treatment, reduction of risk factors, and use of health services. After 1 year, no statistically significant differences were found between the treatment and control groups. The findings indicate that, while home blood pressure monitoring may be useful, it has no measurable short-term impact on these aspects of blood pres...

  4. Transcutaneous oximetry but not arterial toe blood pressure or ankle-brachial index is related to macular thickness in patients with chronic diabetic foot ulcers

    Directory of Open Access Journals (Sweden)

    Per Katzman

    2013-04-01

    Full Text Available Objectives: Diabetes related vascular complications of the lower extremity could be classified as micro- and macrovascular. Ankle-brachial index (ABI and toe blood pressure (TBP measurements evaluate macro-circulation, whereas transcutaneous oximetry (TcPO2 is considered to be a composite measure of vascular function, thereby also reflecting microcirculation. Microvascular disease in the eye involves abnormal capillary permeability and possibly thereby increased macular thickness. The aim of the present clinical study was to evaluate if microvascular disease in the eye was related to measures of foot ischemia in patients with diabetes mellitus. Methods: Twenty consecutive patients with diabetes and chronic full-thickness foot ulcers were included. Peripheral ischemia was diagnosed using TcPO2, TBP and ABI. Macular thickness was measured with optical coherence tomography technique. Results: Based on TcPO2, TBP and ABI measurements 14, 13 and 13 patients, respectively, were classified as ischemic. Patients with ischemic TcPO2 levels at the dorsum of the foot had significantly higher macular thicknesses. This was not the case in patients with ischemic TBP or ABI levels. Conclusion: TcPO2, unlike TBP and ABI, seems to be a clinically relevant measure of peripheral microvascular disease in patients with diabetes mellitus and may, if low, indicate an increased risk of macular edema. [J Exp Integr Med 2013; 3(2.000: 81-85

  5. Diabetes: Monitoring Your Blood Sugar Level

    Science.gov (United States)

    ... to Web version Diabetes | Monitoring Your Blood Sugar Level What tests can I use to check my blood sugar level? There are 2 blood tests that can help ... the past 2-3 months. Testing your A1C level every 3 months is the best way for ...

  6. Gas House Autonomous System Monitoring

    Science.gov (United States)

    Miller, Luke; Edsall, Ashley

    2015-01-01

    Gas House Autonomous System Monitoring (GHASM) will employ Integrated System Health Monitoring (ISHM) of cryogenic fluids in the High Pressure Gas Facility at Stennis Space Center. The preliminary focus of development incorporates the passive monitoring and eventual commanding of the Nitrogen System. ISHM offers generic system awareness, adept at using concepts rather than specific error cases. As an enabler for autonomy, ISHM provides capabilities inclusive of anomaly detection, diagnosis, and abnormality prediction. Advancing ISHM and Autonomous Operation functional capabilities enhances quality of data, optimizes safety, improves cost effectiveness, and has direct benefits to a wide spectrum of aerospace applications.

  7. Blood monitoring systems and methods thereof

    Science.gov (United States)

    Mir, Jose (Inventor); Zander, Dennis (Inventor)

    2012-01-01

    A blood monitoring system is capable of monitoring the blood of a subject in vivo. The blood monitoring system comprises: 1) an array of movable microneedle micromachined within associated wells; 2) array of motion actuators able to move each needle in and out of their associated wells; 3) array of microvalves associated with each microneedle able to control the flow of air around the microneedle; 4) an array of chemical sensors inserted into patient by movable microneedles; 5) an array of inductors able to measure chemical concentration in the vicinity of inserted chemical sensors; 6) conducting vias that provide timed actuating signal signals from a control system to each motion actuator; 7) conducting vias that transmit signal produced by array of chemical sensors to the control system for processing, although the blood monitoring system can comprise other numbers and types of elements in other configurations.

  8. Coronary blood flow dynamics during transcutaneous electrical nerve stimulation for stable angina pectoris associated with severe narrowing of one major coronary artery

    NARCIS (Netherlands)

    Jessurun, GAJ; Tio, RA; De Jongste, MJL; Hautvast, RWM; Den Heijer, P; Crijns, HJGM

    1998-01-01

    To study the effect of transcutaneous electrical nerve stimulation (TENS) on coronary vasomotion, patients with New York Heart Association class III angina pectoris and significant single-vessel left coronary artery disease and who were also scheduled for elective percutaneous transluminal coronary

  9. Blood sugar monitoring with laser diode

    Science.gov (United States)

    Zhang, Xiqin; Chen, Jianhong; Yeo, Joon Hock

    2006-09-01

    In this paper, the non-invasive measurement of blood sugar level was studied by use of near infrared laser diode. The in-vivo experiments were carried out using laser diodes with wavelength 1625nm and 1650nm. Several volunteers were tested before and after drinking glucose solution. We took blood from a fingertip and measured its concentration with a glucose meter while taking signal voltage from laser diode system. The signal voltage was processed by using a computer and blood absorption was obtained. The results show that blood sugar level and blood absorption have similar trends before and after drinking glucose solution. We also compared the trends of drinking glucose solution and pure water and the results show that the difference of blood absorption is obvious. From the results we can see that laser diode is suitable for blood glucose monitoring.

  10. A Novel Infrared Gas Monitor

    Science.gov (United States)

    Wang, Yingding; Zhong, Hongjie

    2000-03-01

    In the paper a novel non-dispersive infrared(IR) gas monitor is described.It is based on the principle that certain gases absorb IR radiation at specific(and often unique) wavelengths.Conventional devices typically include several primary components:a broadband source, usually an incandescent filament,a rotating chopper shutter,a narrow-band filter,a sample tube and a detector. We have developed a number of IR light emitting diodes(LED) having narrow optical bandwidths and which can be intensity modulated by electrical means,for example InAsSbP(4.2 micron)LED.The IR LED can thus replace the thermal source,narrow-band filter and chopper assembly of the conventional IR gas monitor,yielding a solid state,low- powered,compact and almost maintenance-free instrument with high sensitivity and stability and which free of the effects of mechanical vibration too. The detector used in the IR gas monitor is the solid-state detector,such as PbS,PbSe, InSb,HgCdTe,TGS,LT and PZT detector etc. The different configuration of the IR gas monitor is designed.For example,two-path version for measuring methane concentration by monitoring the 3.31 micron absorption band,it can eliminate the interference effects,such as to compensate for LED intensity changes caused by power and temperature variations,and for signal fluctuations due to changes in detector bias. we also have designed portable single-beam version without the sample tube.Its most primary advantage is very cheap(about cost USD 30 ).It measures carbon dioxide concentration by monitoring the 4.25 micron absorption band.Thought its precisions is low,it is used to control carbon dioxide concentration in the air in the green houses and plastic houses(there are about twenty millon one in the China).Because more carbon dioxide will increase the quanity of vegetable and flower production to a greatextent. It also is used in medical,sanitary and antiepidemic applications,such as hospital, store,hotel,cabin and ballroom etc. Key words

  11. Beat-to-Beat Blood Pressure Monitor

    Science.gov (United States)

    Lee, Yong Jin

    2012-01-01

    This device provides non-invasive beat-to-beat blood pressure measurements and can be worn over the upper arm for prolonged durations. Phase and waveform analyses are performed on filtered proximal and distal photoplethysmographic (PPG) waveforms obtained from the brachial artery. The phase analysis is used primarily for the computation of the mean arterial pressure, while the waveform analysis is used primarily to obtain the pulse pressure. Real-time compliance estimate is used to refine both the mean arterial and pulse pressures to provide the beat-to-beat blood pressure measurement. This wearable physiological monitor can be used to continuously observe the beat-to-beat blood pressure (B3P). It can be used to monitor the effect of prolonged exposures to reduced gravitational environments and the effectiveness of various countermeasures. A number of researchers have used pulse wave velocity (PWV) of blood in the arteries to infer the beat-to-beat blood pressure. There has been documentation of relative success, but a device that is able to provide the required accuracy and repeatability has not yet been developed. It has been demonstrated that an accurate and repeatable blood pressure measurement can be obtained by measuring the phase change (e.g., phase velocity), amplitude change, and distortion of the PPG waveforms along the brachial artery. The approach is based on comparing the full PPG waveform between two points along the artery rather than measuring the time-of-flight. Minimizing the measurement separation and confining the measurement area to a single, well-defined artery allows the waveform to retain the general shape between the two measurement points. This allows signal processing of waveforms to determine the phase and amplitude changes. Photoplethysmography, which measures changes in arterial blood volume, is commonly used to obtain heart rate and blood oxygen saturation. The digitized PPG signals are used as inputs into the beat-to-beat blood

  12. 脐血血气分析与胎儿Apgar评分监测缺氧的相关性研究%Study on correlation between umbilical cord blood gas analysis and Apgar scores in monitoring fetal hypoxia

    Institute of Scientific and Technical Information of China (English)

    杨燕

    2013-01-01

    Objective To study the correlation of umbilical cord blood gas analysis and Apgar scores in monitoring fetal hypoxia.Methods 180 cases of bom from March 2011 to March 2012 in our hospital,with the legislation collected cord blood samples for blood gas analysis,and two groups according to the results of fetal Apgar scores,were divided into experimental and control groups.The 90 cases of experimental group had fetal Apgar score ≤7 points,the 90 cases of control group had fetal Apgar score ≥8 points.combining with cord blood research blood gas analysis and fetal Apgar scores,we monitored the two groups of children with hypoxia and statistical analysis.Reaults The experimental group's fetal Apgar score ≤7 Pa02,HC03-,pH values were significantly lower than those of the control group (P<0.05),fetal PaC02 values of the experimental group,BE negative values were significantly higher than those in the control group (P<0.05).Concluaions Umbilical cord blood gas analysis for monitoring fetal Apgrscores neonatal hypoxia.can improve the neonatal diagnostic accuracy,and improve perinatal survival rate.in the control group (P<0.05).Conclusion umbilical cord blood gas analysis to monitor fetal Apgar scores neonatal hypoxia can improve the the neonatal diagnostic accuracy,and improve perinatal survival.%目的 研究脐血血气分析与胎儿Apgar评分监测缺氧的相关性.方法 选择2011年3月至2012年3月我院180例新生儿出生后,立采集脐血血样进行血气分析,并根据胎儿Apgar评分结果,分为实验组和对照组两组.实验组90例胎儿Apgar评分≤7分,对照组90例胎儿Apgar评分≥8分,结合研究脐血血气分析与胎儿Apgar评分情况,监测两组患儿缺氧情况,并进行统计分析.结果 Apgar评分≤7分的实验组胎儿的PaO2、HCO3-、pH值均明显低于Apgar评分≥8分的对照组胎儿(P<0.05),实验组胎儿的PaCO2值、BE负值均明显高于对照组(P<0.05).结论 脐血血气分析与胎

  13. Surgical staple as a transcutaneous transducer for ECG electrodes in burnt skin: safe surgical monitoring in major burns.

    Science.gov (United States)

    Sofos, Stratos S; Tehrani, Hamid; Shokrollahi, Kayvan; James, M I

    2013-06-01

    It is often difficult to apply traditional ECG electrodes on patients with extensive burns due to a large operative site, compromise of sterility, the fact that traditional placement would be within the operative site or because stick-on pads cannot stick due to prep solution, bleeding and other factors. We present an effective solution based on our experience, of using a common staple or "clip" where the ECG electrode is attached. We can see the patient in the prone position with the back having been debrided and grafted. This technical improvisation gives clinicians the ability to monitor safely and accurately the patients' physiological parameters.

  14. Beat-to-Beat Blood Pressure Monitor

    Science.gov (United States)

    Lee, Yong Jin

    2012-01-01

    This device provides non-invasive beat-to-beat blood pressure measurements and can be worn over the upper arm for prolonged durations. Phase and waveform analyses are performed on filtered proximal and distal photoplethysmographic (PPG) waveforms obtained from the brachial artery. The phase analysis is used primarily for the computation of the mean arterial pressure, while the waveform analysis is used primarily to obtain the pulse pressure. Real-time compliance estimate is used to refine both the mean arterial and pulse pressures to provide the beat-to-beat blood pressure measurement. This wearable physiological monitor can be used to continuously observe the beat-to-beat blood pressure (B3P). It can be used to monitor the effect of prolonged exposures to reduced gravitational environments and the effectiveness of various countermeasures. A number of researchers have used pulse wave velocity (PWV) of blood in the arteries to infer the beat-to-beat blood pressure. There has been documentation of relative success, but a device that is able to provide the required accuracy and repeatability has not yet been developed. It has been demonstrated that an accurate and repeatable blood pressure measurement can be obtained by measuring the phase change (e.g., phase velocity), amplitude change, and distortion of the PPG waveforms along the brachial artery. The approach is based on comparing the full PPG waveform between two points along the artery rather than measuring the time-of-flight. Minimizing the measurement separation and confining the measurement area to a single, well-defined artery allows the waveform to retain the general shape between the two measurement points. This allows signal processing of waveforms to determine the phase and amplitude changes.

  15. Blood glucose control and monitoring in the critically ill

    NARCIS (Netherlands)

    R.T.M. van Hooijdonk

    2015-01-01

    This thesis deals with blood glucose control and blood glucose monitoring in intensive care unit (ICU) patients: two important aspects of care for and monitoring of critically ill patients. While the precise targets of blood glucose control in ICU patients remain a matter of debate, currently many,

  16. Managing hypertension with ambulatory blood pressure monitoring.

    Science.gov (United States)

    White, William B; Gulati, Vinay

    2015-02-01

    There has been a dramatic shift in the manner in which blood pressure (BP) is measured to provide far more comprehensive clinical information than that provided by a single set of office BP readings. Extensive clinical and epidemiological research shows an important role of ambulatory BP monitoring (ABPM) in the management of hypertensive patients. A 24-h BP profile helps to determine the absence of nocturnal dipping status and evaluate BP control in patients on antihypertensive therapy. The ability to detect white-coat or masked hypertension is enhanced by ambulatory BP monitoring. In 2001, the Center for Medicare and Medicaid Services approved ABPM for reimbursement for the identification of patients with white-coat hypertension. In 2011, the National Institute for Health and Clinical Excellence (NICE) in the UK published guidelines that recommended the routine use of ABPM in all patients suspected of having hypertension. The European Society of Hypertension (ESH) 2013 guidelines also support greater use of ABPM in clinical practice. While the advantages of ABPM are apparent from a clinical perspective, its use should be considered in relation to its cost, the complexity of data evaluation, as well as patient inconvenience. In this review, we evaluate the clinical importance of ABPM, highlighting its role in the current management of hypertension.

  17. Monitoring gas quality green gas feeding in; Monitoring gaskwaliteit groengasinvoeding

    Energy Technology Data Exchange (ETDEWEB)

    Holstein, J. [DNV KEMA Energy and Sustainability, Arnhem (Netherlands); Polman, E. [Kiwa Technology, Apeldoorn (Netherlands)

    2013-04-15

    Due to the growing number of green gas facilities in the Netherlands more practical knowledge is collected about the production and injection of green gas. Also there was the need to gather data about more practical experiences and knowledge about the gas quality, the performance of gas cleaning and gas treatment systems, as well the integration of green gas in the gas infrastructure. In addition to this, there is a need to get insight in the safety aspects of green gas injection. In order to comply this demand, DNV KEMA en Kiwa Technology measured the quality parameters continuously between June 2012 en January 2013 (three weeks) and discontinuously (gas samples) of green gas at eight production facilities. The measurements have been performed at designated places and are independent from the measurements of the biomethane producer. In order to be sure that the results of DNV KEMA and Kiwa are comparable, a combined measurement program was executed. It results in uniformity for all the measured values: the differences are within the uncertainty level for each component. During the measurement period of three weeks, the gas quality parameters were compared to specifications, written down in the national regulations for the transport and the distribution grid respectively [Dutch] Door het groeiend aantal groengasinvoedingen in Nederland wordt steeds meer praktijkkennis verzameld. Er dient meer praktijkkennis te worden verzameld over de chemische gaskwaliteit, prestaties van de gasreiniging- en gasopwaardering en de wijze van inpassing in de bestaande infrastructuur. Daarnaast is het wenselijk om inzicht te verkrijgen in de veiligheid van groengasinvoeding. Daarop is de groengaskwaliteit op acht locaties over een periode van drie weken continu en discontinu gemeten tussen juni 2012 en januari 2013. De metingen zijn uitgevoerd op een aangewezen plaats door de netbeheerder en staan los van de metingen van de invoeders zelf. Voor het waarborgen van de uniformiteit van

  18. Gas-laser power monitor

    Science.gov (United States)

    Russ, C. E., Jr.

    1981-01-01

    Device attaches simply to front of laser housing for continuous monitoring of power output. Monitor is calibrated to read either total output or power generated in test volume. It is fabricated from four black-anodized aluminum parts; crown glass positioned at Brewster angle reflects 0.33 percent of beam onto photodiode calibrated for electrical output proportional to laser power. Unlike conventional calorimeter, monitor does not interrupt laser beams, and fast-response diode allows instantaneous tracking of power fluctuations.

  19. 20 CFR 718.105 - Arterial blood-gas studies.

    Science.gov (United States)

    2010-04-01

    ...) Duration and type of exercise; (8) Pulse rate at the time the blood sample was drawn; (9) Time between... at rest or during exercise. No blood-gas study shall be performed if medically contraindicated. (b) A... blood-gas test at rest do not satisfy the requirements of Appendix C to this part, an exercise...

  20. Prediction of arterial blood gas values from arterialized earlobe blood gas values in patients treated with mechanical ventilation

    Directory of Open Access Journals (Sweden)

    Honarmand Azim

    2008-01-01

    earlobe blood gas can accurately predict the ABG values of pH, PCO 2 , BE, and HCO 3 - for patients who do not require regular continuous blood pressure measurements and close monitoring of arterial PO 2 measurements.

  1. Greenhouse gas emissions studies using a portable TDL gas monitor

    International Nuclear Information System (INIS)

    The Boreal Laser company developed a portable open path gas monitor called the GasFinder. The GasFinder is based on laser technology and is an ideal tool for studying greenhouse gas emissions. The response time is one second, and the GasFinder is specific to the gas under study. In addition, it has a broad dynamic range. Very small and light, weighing only 5 kilograms, the GasFinder is permanently calibrated. The alignment of the GasFinder is easy and stable. The path length has no effect on response, enabling the GasFinder to provide a series of paths measurements in quick succession for different lengths (between 1 metre and 1000 metres). The establishment of baseline emissions and the determination of the effectiveness of process improvements for the reduction of emissions are instances where the GasFinder has already been successfully used. The authors discussed some examples concerning the detection of methane leak in natural gas production, processing and transportation and carbon dioxide measurements from both natural and man-made sources

  2. Ionization chamber for monitoring radioactive gas

    International Nuclear Information System (INIS)

    This present invention provides simple, effective and accurate cumulative measurement of radioactive gas over a time period. Measurements of radioactive gas are important for many purposes. Tritium concentrations in potentially exposed workers are measured, for example, with periodic urine specimens. Carbon-14 serves as a useful research tool for monitoring the progress of many chemical and biological reactions and interactions. For example, many microorganisms break down carbon-14 containing compounds in sugar to produce carbon-14 dioxide gas which can be collected and measured to determine various characteristics of the microorganisms. Both tritium and carbon-14 dioxide produce low energy radiation which cannot be easily measured by conventional radioactivity detectors. (author). 4 figs

  3. Wearable Beat to Beat Blood Pressure Monitor Project

    Data.gov (United States)

    National Aeronautics and Space Administration — A key component of NASA's human exploration programs is a system that monitors the health of the crew during space missions. The wearable beat-to-beat blood...

  4. Ambulatory Blood Pressure Monitoring and Circadian Rhythm of Blood Pressure in Diabetes Mellitus

    OpenAIRE

    Elena Matteucci; Ottavio Giampietro

    2013-01-01

    Systolic and diastolic blood pressures display a circadian rhythmicity that can be assessed by 24-hour ambulatory blood pressure monitoring and analysed using the cosinor procedure. Altered characteristics to the circadian rhythm of blood pressure, which may result in adverse health outcomes, have been observed in both prediabetes and diabetes. We have investigated the circadian variability of blood pressure in patients with type 1 and type 2 diabetes. Chronobiologically interpreted ambulator...

  5. Arterialisation of transcutaneous oxygen and carbon dioxide.

    OpenAIRE

    Broadhurst, E; Helms, P; Vyas, H; Cheriyan, G

    1988-01-01

    We compared previously calculated global correction factors for oxygen and carbon dioxide arterial/transcutaneous ratios with individual in vivo calibrations from the first arterial sample. In infants beyond the neonatal period and older children in vivo calibration confers little benefit over the use of a global calibration correction factor for transcutaneous carbon dioxide, and may reduce the precision with which arterial oxygen can be estimated from transcutaneous oxygen.

  6. New Approaches to Evaluating and Monitoring Blood Pressure.

    Science.gov (United States)

    Goldberg, Elizabeth M; Levy, Phillip D

    2016-06-01

    Digital health innovations for hypertension include cuffless blood pressure sensors, wireless smartphone-enabled upper arm blood pressure monitors, mobile applications, and remote monitoring technologies. Wearable trackers have drawn interest from medical professionals and patients alike. They have the potential to improve hypertension control and medication adherence through easier logging of repeated blood pressure measurements, better connectivity with health-care providers, and medication reminder alerts. With increasing emphasis on home and ambulatory blood pressure monitoring to confirm hypertension prior to treatment, such devices also can help improve the diagnostic landscape. However, privacy, accuracy, and cost concerns have prevented widespread clinical uptake. To foster implementation, device designers and clinical researchers should collaborate on development of rigorous clinical trials that test cardiovascular outcomes associated with emerging technologies. We review the current literature on mobile health technologies and novel diagnostic and management protocols and make recommendations on how to incorporate these innovations into practice. PMID:27137524

  7. Dependence of transcutaneous oxygen tension on local arteriovenous pressure gradient in normal subjects.

    Science.gov (United States)

    Wyss, C R; Matsen, F A; King, R V; Simmons, C W; Burgess, E M

    1981-05-01

    1. We studied the relationship between transcutaneous oxygen tension at the foot and local arteriovenous pressure difference in 15 normal men and women; arteriovenous pressure difference was varied by changing the height of the foot with respect to the heart and by applying external pressure to the foot. 2. Control transcutaneous oxygen tension was 67 +/- 9 SD mmHg (8.9 +/- 1.2 kPa) at a control arteriovenous pressure difference of 80 +/- 6 SD mmHg (10.6 +/- 0.8 kPa). 3. In every subject transcutaneous oxygen tension fell non-linearly with a decrease in arteriovenous pressure difference; transcutaneous oxygen tension was relatively insensitive to changes in arteriovenous pressure difference when arteriovenous pressure difference was high, but always fell sharply to zero at some positive arteriovenous pressure difference [range 13-34 mmHg (1.7-4.5 kPa)]. 4. An analysis of the data indicated that transcutaneous oxygen tension varied with arteriovenous pressure difference approximately as the oxygen tension of cutaneous venous blood under the sensor varied (in the absence of changes in cutaneous vascular resistance and oxygen consumption). 5. This analysis was supported by studies in three subjects in whom the oxygen tension of superficial venous drainage from a warmed hand or foot was measured along with transcutaneous oxygen tension while arteriovenous pressure difference was varied.

  8. A xenon gas purity monitor for EXO

    CERN Document Server

    Dobi, A; Herrin, S; Odian, A; Prescott, C Y; Rowson, P C; Ackerman, N; Aharmin, B; Auger, M; Barbeau, P S; Barry, K; Benitez-Medina, C; Breidenbach, M; Cook, S; Counts, I; Daniels, T; DeVoe, R; Dolinski, M J; Donato, K; Fairbank, W; Farine, J; Giroux, G; Gornea, R; Graham, K; Gratta, G; Green, M; Hagemann, C; Hall, K; Hallman, D; Hargrove, C; Karelin, A; Kaufman, L J; Kuchenkov, A; Kumar, K; Lacey, J; Leonard, D S; LePort, F; Mackay, D; MacLellan, R; Mong, B; Diez, M Montero; Muller, A R; Neilson, R; Niner, E; O'Sullivan, K; Piepke, A; Pocar, A; Pushkin, K; Rollin, E; Sinclair, D; Slutsky, S; Stekhanov, V; Twelker, K; Voskanian, N; Vuilleumier, J -L; Wichoski, U; Wodin, J; Yang, L; Yen, Y -R

    2011-01-01

    We discuss the design, operation, and calibration of two versions of a xenon gas purity monitor (GPM) developed for the EXO double beta decay program. The devices are sensitive to concentrations of oxygen well below 1 ppb at an ambient gas pressure of one atmosphere or more. The theory of operation of the GPM is discussed along with the interactions of oxygen and other impurities with the GPM's tungsten filament. Lab tests and experiences in commissioning the EXO-200 double beta decay experiment are described. These devices can also be used on other noble gases.

  9. Cuff inflation during ambulatory blood pressure monitoring and heart rate

    Directory of Open Access Journals (Sweden)

    Mia Skov-Madsen

    2008-11-01

    Full Text Available Mia Skov-Madsen, My Svensson, Jeppe Hagstrup ChristensenDepartment of Nephrology, Aarhus University Hospital, Aalborg, DenmarkIntroduction: Twenty four-hour ambulatory blood pressure monitoring is a clinically validated procedure in evaluation of blood pressure (BP. We hypothesised that the discomfort during cuff inflation would increase the heart rate (HR measured with 24-h ambulatory BP monitoring compared to a following HR measurement with a 24-h Holter monitor.Methods: The study population (n = 56 were recruited from the outpatient’s clinic at the Department of Nephrology, Aalborg Hospital, Aarhus University Hospital at Aalborg, Denmark. All the patients had chronic kidney disease (CKD. We compared HR measured with a 24-h Holter monitor with a following HR measured by a 24-h ambulatory BP monitoring.Results: We found a highly significant correlation between the HR measured with the Holter monitor and HR measured with 24-h ambulatory blood pressure monitoring (r = 0.77, p < 0.001. Using the Bland-Altman plot, the mean difference in HR was only 0.5 beat/min during 24 hours with acceptable limits of agreement for both high and low HR levels. Dividing the patients into groups according to betablocker treatment, body mass index, age, sex, angiotensin-converting enzyme inhibitor treatment, statins treatment, diuretic treatment, or calcium channel blocker treatment revealed similar results as described above.Conclusion: The results indicate that the discomfort induced by cuff inflation during 24-h ambulatory BP monitoring does not increase HR. Thus, 24-h ambulatory BP monitoring may be a reliable measurement of the BP among people with CKD.Keywords: ambulatory blood pressure monitoring, Holter monitoring, heart rate, chronic kidney disease, hypertension

  10. Apparatus and method for monitoring of gas having stable isotopes

    Science.gov (United States)

    Clegg, Samuel M; Fessenden-Rahn, Julianna E

    2013-03-05

    Gas having stable isotopes is monitored continuously by using a system that sends a modulated laser beam to the gas and collects and transmits the light not absorbed by the gas to a detector. Gas from geological storage, or from the atmosphere can be monitored continuously without collecting samples and transporting them to a lab.

  11. Noninvasive 24-hour ambulatory arterial blood pressure monitoring in cirrhosis

    DEFF Research Database (Denmark)

    Møller, Søren; Wiinberg, N; Henriksen, Jens Henrik

    1995-01-01

    device for monitoring blood pressure in 35 patients with cirrhosis and 35 healthy matched controls. During the daytime, SBP, DBP, and MAP were significantly lower in the patients than in the controls (median 118 vs. 127; 70 vs. 78; 86 vs. 94 mm Hg, P blood pressures......Cirrhotic patients have disturbed systemic hemodynamics with reduced arterial blood pressure, but this has not been investigated during daily activity and sleep. Systolic (SBP), diastolic (DBP), and mean arterial blood pressure (MAP), and heart rate (HR) were measured by an automatic ambulant...... were almost similar in the two groups (108 vs. 110; 65 vs. 67; 78 vs. 82 mm Hg, NS). Conversely, HR was significantly higher in the patients both in the daytime (86 vs. 72/min, P blood pressure and HR from daytime...

  12. Inherently safe passive gas monitoring system

    Energy Technology Data Exchange (ETDEWEB)

    Cordaro, Joseph V.; Bellamy, John Stephen; Shuler, James M.; Shull, Davis J.; Leduc, Daniel R.

    2016-09-06

    Generally, the present disclosure is directed to gas monitoring systems that use inductive power transfer to safely power an electrically passive device included within a nuclear material storage container. In particular, the electrically passive device can include an inductive power receiver for receiving inductive power transfer through a wall of the nuclear material storage container. The power received by the inductive power receiver can be used to power one or more sensors included in the device. Thus, the device is not required to include active power generation components such as, for example, a battery, that increase the risk of a spark igniting flammable gases within the container.

  13. Predictors of Daily Blood Glucose Monitoring in Appalachian Ohio

    Science.gov (United States)

    Raffle, Holly; Ware, Lezlee J.; Ruhil, Anirudh V. S.; Hamel-Lambert, Jane; Denham, Sharon A.

    2012-01-01

    Objective: To determine factors contributing to successful diabetes self-management in Appalachia, as evidenced by daily blood glucose monitoring. Methods: A telephone survey (N = 3841) was conducted to assess health status and health care access. The current investigation is limited to the subset of this sample who report having diabetes (N =…

  14. Blood perfusion and pH monitoring in organs by laser-induced fluorescence spectroscopy

    Science.gov (United States)

    Vari, Sandor G.; Papazoglou, Theodore G.; Pergadia, Vani R.; Stavridi, Marigo; Snyder, Wendy J.; Papaioannou, Thanassis; Duffy, J. T.; Weiss, Andrew B.; Thomas, Reem; Grundfest, Warren S.

    1994-01-01

    Sensitivity of laser-induced fluorescence spectroscopy (LIFS) in detecting a change in tissue pH, and blood perfusion was determined. Rabbits were anesthetized, paralyzed, and mechanically ventilated. The arterial and venous blood supplies of the kidney were isolated and ligated to alter the perfusion. The femoral artery was cannulated to extract samples for blood gas analysis. A 308-nm XeCl was used as an excitation source. A 600 micrometers core diameter fiber was used for fluorescence acquisition, and the spectra analyzed by an optical multichannel analyzer (EG & G, OMA III). the corresponding intensity ratio R equals INADH / ICOLL was used as an index for respiratory acidosis. Blood perfusion was assessed using the following algorithm: (IELAS minus ICOLL) divided by (INADH minus ICOLL). The intensity ratio linearly decreased with the reduction of blood perfusion. When we totally occluded the artery the ratio decreased tenfold when compared to the ratio of a fully perfused kidney. Results of monitoring blood acidosis by laser-induced fluorescence spectroscopy shows a significant trend between pH and intensity ratio. Since all the slopes were negative, there is an obvious significant correlation between the pH and NADH.COLLAGEN RATIO. Blue-light-induced fluorescence measurements and ratio fluorometry is a sensitive method for monitoring blood perfusion and acidity or alkalinity of an organ.

  15. Photoacoustic monitoring and imaging of blood vessels in tissue

    Science.gov (United States)

    Kolkman, Roy G. M.; Pilatou, Magdalena C.; Steenbergen, Wiendelt; de Mul, Frits F. M.

    2002-06-01

    Using very sensitive photoacoustical detectors we localized and monitored the blood content in tissue. In these detectors a PVdF-layer has been used as piezo-electric material and also fibers for the illumination of the sample are integrated. The resolution is about 20micrometers in depth and about 50-100micrometers laterally. The wavelengths of the laser light were 532 and 1064 nm. With these colors we can measure at different depths in tissue. The measurements concerned blood perfusion in real tissue: vessels in chicken breast, in test animals at various positions and in the human arm.

  16. TRANSCUTANEOUS DRUG DELIVERY SYSTEM: A COMPREHENSIVE REVIEW

    OpenAIRE

    Sandhu Premjeet; Kataria Sahil; Bilandi Ajay; Jain Sonam; Rathore Devashish

    2011-01-01

    Conventional drug delivery systems are often not suitable for new protein based and other Therapeutic compounds produced by modern technology. Therefore an alternative Approach to deliver these drugs can be achieved through the skin in the form of transcutaneous drug delivery system. Modern medicine has responded with the development of methods to deliver drug transcutanously (through) the skin for therapeutic use as an alternative to traditional route including oral, intravascular, intramusc...

  17. Novel optoacoustic array for noninvasive monitoring of blood parameters

    Science.gov (United States)

    Andreev, Valeriy G.; Petrov, Yuriy Y.; Prough, Donald S.; Petrova, Irina Y.; Esenaliev, Rinat O.

    2009-02-01

    Noninvasive monitoring of blood parameters such as total hemoglobin concentration and saturation (oxygenation) is important for diagnostics in large populations of patients. We developed a novel optoacoustic array for monitoring of these variables in arteries and veins. The array allows for measurements without scanning, reduces the data acquisition time, and minimizes the influence of motion artifacts. The array combines a custom-made fiber-optic delivery system and multiple piezoelectric transducers. We tested the array in vitro and in vivo. The array design and materials allowed for sensitive measurement of optoacoustic signals without distortion and provided real time measurement of these parameters both in vitro and in vivo without scanning.

  18. Oscillometric continuous blood pressure sensing for wearable health monitoring system

    CERN Document Server

    Gelao, Gennaro; Passaro, Vittorio M N; Perri, Anna Gina

    2015-01-01

    In this paper we present an acquisition chain for the measurement of blood arterial pressure based on the oscillometric method. This method does not suffer from any limitation as the well-known auscultatory method and it is suited for wearable health monitoring systems. The device uses a pressure sensor whose signal is filtered, digitalized and analyzed by a microcontroller. Local analysis allows the evaluation of the systolic and diastolic pressure values which can be used for local alarms, data collection and remote monitoring.

  19. Self-monitoring ultrasonic gas flow meter

    Science.gov (United States)

    Lin, Y.; Hans, V.

    2005-01-01

    Ultrasound is predestined for gas flow velocity measurements on account of its high sensitivity to all kinds of natural and artificial turbulences in the fluid. Vortex measurements behind a bluff body as well as cross-correlation methods have been proved good. Cross-correlation measurements of natural structures determine the most frequent velocity components in the fluid. Therefore, the measured flow velocity deviates from the real mean flow velocity because of a skewed probability density distribution of the velocity components. Vortex measurements base on the principle that the frequency of the vortices generated in the wake of a bluff body is proportional to the mean flow velocity. The measurement of the periodic vortices with cross-correlation functions leads to the direct determination of the real mean flow velocity. The combination of both measuring methods results in a self-monitoring system.

  20. TRANSCUTANEOUS DRUG DELIVERY SYSTEM: A COMPREHENSIVE REVIEW

    Directory of Open Access Journals (Sweden)

    Sandhu Premjeet

    2011-12-01

    Full Text Available Conventional drug delivery systems are often not suitable for new protein based and other Therapeutic compounds produced by modern technology. Therefore an alternative Approach to deliver these drugs can be achieved through the skin in the form of transcutaneous drug delivery system. Modern medicine has responded with the development of methods to deliver drug transcutanously (through the skin for therapeutic use as an alternative to traditional route including oral, intravascular, intramuscular, subcutaneous, and sublingual. Transcutaneous drug delivery has many theoretic and practical advantage and disadvantages, and such issues are often a concern for both clinicians and patients. Transcutaneous patches are flexible pharmaceutical preparations of varying sizes, containing one or more active ingredient, intended to be applied to the unbroken skin in order to deliver the active ingredient to the systemic circulation after passing through the skin barriers. A Transcutaneous patch or skin patch is a medicated adhesive patch that is placed on the skin to deliver a specific dose of medication through the skin and into the bloodstream. Often, this promotes healing to an injured area of the body. In this method, the drug enters the bloodstream directly through skin and it avoid first pass effect. Characterization of Transcutaneous patch are necessary because check it’s quality, size, time of onset & duration, adhesive property, thickness, weight of patch, moisture of content, uniformity & cutaneous toxicological studies. Their requirements for evaluation are HPLC, U.V. spectrophotometer, screw gauge, digital balance, desiccators, thin layer chromatography & K.C. Cell used.

  1. Noble gas atmospheric monitoring at reprocessing facilities

    Energy Technology Data Exchange (ETDEWEB)

    Nakhleh, C.W.; Perry, R.T. Jr.; Poths, J.; Stanbro, W.D.; Wilson, W.B.; Fearey, B.L.

    1997-05-01

    The discovery in Iraq after the Gulf War of the existence of a large clandestine nuclear-weapon program has led to an across-the-board international effort, dubbed Programme 93+2, to improve the effectiveness and efficiency of International Atomic Energy Agency (IAEA) safeguards. One particularly significant potential change is the introduction of environmental monitoring (EM) techniques as an adjunct to traditional safeguards methods. Monitoring of stable noble gas (Kr, Xe) isotopic abundances at reprocessing plant stacks appears to be able to yield information on the burnup and type of the fuel being processed. To estimate the size of these signals, model calculations of the production of stable Kr, Xe nuclides in reactor fuel and the subsequent dilution of these nuclides in the plant stack are carried out for two case studies: reprocessing of PWR fuel with a burnup of 35 GWd/tU, and reprocessing of CAND fuel with a burnup of 1 GWd/tU. For each case, a maximum-likelihood analysis is used to determine the fuel burnup and type from the isotopic data.

  2. Laser Doppler perfusion monitoring and imaging of blood microcirculation

    Science.gov (United States)

    Nilsson, Gert E.; Wardell, Karin

    1994-07-01

    Laser Doppler perfusion monitoring is a method of assessing tissue perfusion based on measurements performed using Doppler broadening of monochromatic light scattered in moving blood cells. Ever since laser Doppler perfusion monitors became available about 15 years ago they have been used in numerous applications in both clinical and laboratory settings. The high spatial resolution has in practice manifested itself as one of the main limitations of the method. The reason for this is the difficulty in attaining reproducible values at successive measurement sites because most skin tissue possesses a substantial variation in blood flow even at adjacent measurement sites. In order to overcome this difficulty the laser Doppler perfusion imager was developed. In this camera-like device, the laser beam successively scans the tissue and the Doppler components of the backscattered light are detected by a remote photodiode. After a scanning procedure is complete, a color-coded perfusion map showing the spatial variation of skin blood flow is displayed on a monitor. The operating principle and early applications of this emerging technology are addressed in further detail.

  3. Noninvasive optoacoustic monitoring of cerebral venous blood oxygenation in newborns

    Science.gov (United States)

    Petrov, Irene Y.; Wynne, Karon E.; Petrov, Yuriy; Esenaliev, Rinat O.; Richardson, C. Joan; Prough, Donald S.

    2012-02-01

    Cerebral ischemia after birth and during labor is a major cause of death and severe complications such as cerebral palsy. In the USA alone, cerebral palsy results in permanent disability of 10,000 newborns per year and approximately 500,000 of the total population. Currently, no technology is capable of direct monitoring of cerebral oxygenation in newborns. This study proposes the use of an optoacoustic technique for noninvasive cerebral ischemia monitoring by probing the superior sagittal sinus (SSS), a large central cerebral vein. We developed and built a multi-wavelength, near-infrared optoacoustic system suitable for noninvasive monitoring of cerebral ischemia in newborns with normal weight (NBW), low birth-weight (LBW, 1500 - 2499 g) and very low birth-weight (VLBW, cerebral oxygenation would suggest that no therapy was necessary; conversely, evidence of cerebral ischemia would prompt therapy to increase cerebral blood flow.

  4. Development of blood extraction system for health monitoring system

    Science.gov (United States)

    Tsuchiya, Kazuyoshi; Nakanishi, Naoyuki; Nakamachi, Eiji

    2004-03-01

    The purpose of this research is to develop the compact human blood sampling device applied for a health monitoring system(HMS), which is called "Mobile Hospital". The HMS consists of (1) a micro electrical pumping system for blood extraction, (2) a bio-sensor to detect and evaluate an amount of Glucose, Cholesterol and Urea in extracted blood, by using enzyme such as Glucoseoxidase (GOD), Cholesteroloxidase and Urease. The mechanical design elements of the device are bio-compatible microneedle, indentation unit using a shape memory alloy(SMA) actuator and pumping unit using a piezoelectric microactuator. The design concept is the biomimetic micromachine of female mosquito"s blood sampling mechanism. The performances of the main mechanical elements such as indentation force of the microneedle, actual stroke of the indentation unit using a SMA actuator and liquid sampling ability of the pumping unit using PZT piezoelectric microactuator were measured. The 3 mm stroke of the indentation load generated by SMA actuator was 0.8mN. The amount of imitation blood extracted by using bimorph PZT actuators was about 0.5 microliters for 10 sec. A 60-micrometer outer diameter and 25-micrometer inner diameter Titanium microneedle, which size is same as female mosquito"s labium, was produced by sputter deposition.

  5. Ambulatory blood pressure monitoring: coming of age in nephrology.

    Science.gov (United States)

    Townsend, R R; Ford, V

    1996-11-01

    The number of patients undergoing ambulatory blood pressure monitoring (ABPM) and the number of publications using this technique to evaluate the risks and effects of high blood pressure on target organs has been increasing, and dramatically so, in the last 5 years. Much of this growth has centered on the role of the blood pressure load (the percentage of systolic or diastolic readings above a preset value during a specific time period) and the changes in blood pressures levels that occur, with sleep. Although many studies are focused on the interaction between blood pressure (as assessed by ABPM) and the heart, interest is growing in the application of ABPM to the practice of nephrology. This paper discusses some of the technical aspects of ABPM, followed by a review of five areas of clinical research using ABPM, and which are relevant to renal medicine: microalbuminuria, renal function, renovascular hypertension, dialysis (hemodialysis and continuous ambulatory peritoneal dialysis), and transplantation. Despite a general lack of reimbursement for performance of the ABPM procedure, the growth in its usage and the willingness of clinicians to withhold or alter therapy on the basis of ABPM readings is testimony to its clinical value in the management of hypertension. PMID:8959618

  6. Performance of a coincidence based blood activity monitor

    International Nuclear Information System (INIS)

    A new device has been constructed that measures the positron emitting radio-tracer concentration in arterial blood by extracting blood with a peristaltic pump, then measuring the activity concentration by detecting coincident pairs of 511 keV photons with a pair of heavy inorganic scintillators attached to photomultiplier tubes. The sensitivity of this device is experimentally determined to be 610 counts/second per μCi/ml, and has a paralyzing dead time of 1.2 μs, so is capable of measuring blood activity concentration as high as 1 mCi/ml. Its performance is compared to two other blood monitoring methods: discrete blood samples counted with a well counter and device that uses a plastic scintillator to directly detect positrons. The positron detection efficiency of this device for 18F is greater than the plastic scintillation counter, and also eliminates the radioisotope dependent correction factors necessary to convert count rate to absolute concentration. Coincident photon detection also has the potential of reducing the background compared to direct positron detection, thereby increasing the minimum detectable isotope concentration. 10 refs., 6 figs

  7. [A gas chromatographic method for determining acetaldehyde in cadaver blood].

    Science.gov (United States)

    Savich, V I; Valladares, Kh A; Gusakov, Iu A; Skachko, Z M

    1990-01-01

    Gas-chromatographic method of acetaldehyde detection in blood of subjects who died of alcoholic intoxication is suggested. Method is simple, does not require additional expenses, can be readily used in medicolegal practice and in difficult cases it may help the expert to make an objective conclusion on the cause of death. PMID:2087747

  8. 21 CFR 870.4410 - Cardiopulmonary bypass in-line blood gas sensor.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Cardiopulmonary bypass in-line blood gas sensor... Cardiopulmonary bypass in-line blood gas sensor. (a) Identification. A cardiopulmonary bypass in-line blood gas sensor is a transducer that measures the level of gases in the blood. (b) Classification. Class...

  9. Microseism Monitoring System for Coal and Gas Outburst

    Directory of Open Access Journals (Sweden)

    Li Zhenbi

    2012-09-01

    Full Text Available The outburst forecast of coal and gas is a complex system engineering. On the basis of the analysis of microseism monitoring principle, a simplex positioning algorithm for microseism monitoring is designed; a mine microseism monitoring system is established to canalize mine microseism. Mechanism of the error producing and noise reduction measures is studied. We can analyze the data of the microseism monitoring to find coal or rock vibration caused by mining activities. Microseism monitoring system can capture real-time positioning information. It also can timely, accurately monitor and position these microseism events and the mining microseism event, which provide the pressure monitoring, the prediction of gas outstanding and the next step gas coal bed mining monitoring with reference experience.

  10. Microseism Monitoring System for Coal and Gas Outburst

    OpenAIRE

    Li Zhenbi; Zhao Baiting

    2012-01-01

    The outburst forecast of coal and gas is a complex system engineering. On the basis of the analysis of microseism monitoring principle, a simplex positioning algorithm for microseism monitoring is designed; a mine microseism monitoring system is established to canalize mine microseism. Mechanism of the error producing and noise reduction measures is studied. We can analyze the data of the microseism monitoring to find coal or rock vibration caused by mining activities. Microseism monitoring s...

  11. Optical coherence tomography for blood glucose monitoring through signal attenuation

    Science.gov (United States)

    De Pretto, Lucas R.; Yoshimura, Tania M.; Ribeiro, Martha S.; de Freitas, Anderson Z.

    2016-03-01

    Development of non-invasive techniques for glucose monitoring is crucial to improve glucose control and treatment adherence in patients with diabetes. Hereafter, Optical Coherence Tomography (OCT) may offer a good alternative for portable glucometers, since it uses light to probe samples. Changes in the object of interest can alter the intensity of light returning from the sample and, through it, one can estimate the sample's attenuation coefficient (μt) of light. In this work, we aimed to explore the behavior of μt of mouse's blood under increasing glucose concentrations. Different samples were prepared in four glucose concentrations using a mixture of heparinized blood, phosphate buffer saline and glucose. Blood glucose concentrations were measured with a blood glucometer, for reference. We have also prepared other samples diluting the blood in isotonic saline solution to check the effect of a higher multiple-scattering component on the ability of the technique to differentiate glucose levels based on μt. The OCT system used was a commercial Spectral Radar OCT with 930 nm central wavelength and spectral bandwidth (FWHM) of 100 nm. The system proved to be sensitive for all blood glucose concentrations tested, with good correlations with the obtained attenuation coefficients. A linear tendency was observed, with an increase in attenuation with higher values of glucose. Statistical difference was observed between all groups (p<0.001). This work opens the possibility towards a non-invasive diagnostic modality using OCT for glycemic control, which eliminates the use of analytes and/or test strips, as in the case with commercially available glucometers.

  12. Transcutaneous measurement of the arterial input function in positron emission tomography

    International Nuclear Information System (INIS)

    Positron emission tomography (PET) provides a powerful tool in medical research. Biochemical function can be both precisely localized and quantitatively measured. To achieve reliable quantitation it is necessary to know the time course of activity concentration in the arterial blood during the measurement. In this study the arterial blood curve from the brachial artery is compared to the activity measured in the internal carotid artery with a new transcutaneous detector

  13. High-resolution ultrasound imaging and noninvasive optoacoustic monitoring of blood variables in peripheral blood vessels

    Science.gov (United States)

    Petrov, Irene Y.; Petrov, Yuriy; Prough, Donald S.; Esenaliev, Rinat O.

    2011-03-01

    Ultrasound imaging is being widely used in clinics to obtain diagnostic information non-invasively and in real time. A high-resolution ultrasound imaging platform, Vevo (VisualSonics, Inc.) provides in vivo, real-time images with exceptional resolution (up to 30 microns) using high-frequency transducers (up to 80 MHz). Recently, we built optoacoustic systems for probing radial artery and peripheral veins that can be used for noninvasive monitoring of total hemoglobin concentration, oxyhemoglobin saturation, and concentration of important endogenous and exogenous chromophores (such as ICG). In this work we used the high-resolution ultrasound imaging system Vevo 770 for visualization of the radial artery and peripheral veins and acquired corresponding optoacoustic signals from them using the optoacoustic systems. Analysis of the optoacoustic data with a specially developed algorithm allowed for measurement of blood oxygenation in the blood vessels as well as for continuous, real-time monitoring of arterial and venous blood oxygenation. Our results indicate that: 1) the optoacoustic technique (unlike pure optical approaches and other noninvasive techniques) is capable of accurate peripheral venous oxygenation measurement; and 2) peripheral venous oxygenation is dependent on skin temperature and local hemodynamics. Moreover, we performed for the first time (to the best of our knowledge) a comparative study of optoacoustic arterial oximetry and a standard pulse oximeter in humans and demonstrated superior performance of the optoacoustic arterial oximeter, in particular at low blood flow.

  14. Ambulatory Blood Pressure Monitoring – Clinical Practice Recommendations

    Directory of Open Access Journals (Sweden)

    Katalin Mako

    2016-09-01

    Full Text Available Ambulatory blood pressure monitoring (ABPM became a subject of considerable scientific interest. Due to the increasing use of the ABPM in everyday clinical practice it is important that all the users have a correct knowledge on the clinical indications, the methodology of using the device including some technical issues and the interpretation of results. In the last years several guidelines and position papers have been published with recommendations for the monitoring process, reference values, for clinical practice and research. This paper represents a summary of the most important aspects related to the use of ABPM in daily practice, being a synthesis of recommendations from the recent published guidelines and position papers. This reference article presents the practical and technical issues of ABPM, the use of this method in special situations, the clinical interpretation of measured values including the presentation of different ABPM patterns, derived parameters, the prognostic significance and the limitations of this method.

  15. Transcutaneous Bilirubin Measurement in Preterm Neonates

    Directory of Open Access Journals (Sweden)

    Paymaneh Alizadeh Taheri

    2012-11-01

    Full Text Available Hyperbilirubinemia is a common problem during neonatal period especially in preterm neonates. Transcutaneous bilirubin measurement (TcB by special devices had been documented as an effective tool for predicting neonatal jaundice in full term neonate, but for preterm infants the present evidences are controversial. We carried out a prospective study in Shariati Hospital NICU. 126 paired TcB/total serum bilirubin (TSB measurements were obtained. TcB (on forehead and sternum were measured using JH2-1A device for every admitted preterm infant who clinically showed jaundice and TSB measurements was obtained within 30 min of TcB. 58 (46% were male and 68 (54% were female. The mean gestational age was 31 week and mean birth weight was 1728 ± 60 g. 30 percent of neonates were ill. The mean value obtained by TBS was 8.8 mg/dl and for frontal TcB was 8.2mg/dl and for sternal TcB was 7.4mg/dl. There were good correlation between TBS and TcB and the maximum correlation were seen in 33-37 weeks of gestation and birth weight more than 2500 g with forehead TcB measurement. Healthy preterm infants had significant correlation of TSB and TcB (r=0.56, P<0.001 and ill preterm neonate had r =0.82, P<0.001. The correlation between TBS and TcB with and without phototherapy was r=0.66, P=0.000 and r=0.69, P=0.000 respectively. Although TcB measurement may underestimate TBS but there is significant correlation between TcB and TBS in preterm cases even in ill neonate or who receiving phototherapy. This method can be used for determination of bilirbin level in preterm neonate and reduces the number of blood sampling.

  16. Ultrasonic monitoring of droplets' evaporation: Application to human whole blood.

    Science.gov (United States)

    Laux, D; Ferrandis, J Y; Brutin, D

    2016-09-01

    During a colloidal droplet evaporation, a sol-gel transition can be observed and is described by the desiccation time τD and the gelation time τG. These characteristic times, which can be linked to viscoelastic properties of the droplet and to its composition, are classically rated by analysis of mass droplet evolution during evaporation. Even if monitoring mass evolution versus time seems straightforward, this approach is very sensitive to environmental conditions (vibrations, air flow…) as mass has to be evaluated very accurately using ultra-sensitive weighing scales. In this study we investigated the potentialities of ultrasonic shear reflectometry to assess τD and τG in a simple and reliable manner. In order to validate this approach, our study has focused on blood droplets evaporation on which a great deal of work has recently been published. Desiccation and gelation times measured with shear ultrasonic reflectometry have been perfectly correlated to values obtained from mass versus time analysis. This ultrasonic method which is not very sensitive to environmental perturbations is therefore very interesting to monitor the drying of blood droplets in a simple manner and is more generally suitable for complex fluid droplets evaporation investigation.

  17. Ultrasonic monitoring of droplets' evaporation: Application to human whole blood.

    Science.gov (United States)

    Laux, D; Ferrandis, J Y; Brutin, D

    2016-09-01

    During a colloidal droplet evaporation, a sol-gel transition can be observed and is described by the desiccation time τD and the gelation time τG. These characteristic times, which can be linked to viscoelastic properties of the droplet and to its composition, are classically rated by analysis of mass droplet evolution during evaporation. Even if monitoring mass evolution versus time seems straightforward, this approach is very sensitive to environmental conditions (vibrations, air flow…) as mass has to be evaluated very accurately using ultra-sensitive weighing scales. In this study we investigated the potentialities of ultrasonic shear reflectometry to assess τD and τG in a simple and reliable manner. In order to validate this approach, our study has focused on blood droplets evaporation on which a great deal of work has recently been published. Desiccation and gelation times measured with shear ultrasonic reflectometry have been perfectly correlated to values obtained from mass versus time analysis. This ultrasonic method which is not very sensitive to environmental perturbations is therefore very interesting to monitor the drying of blood droplets in a simple manner and is more generally suitable for complex fluid droplets evaporation investigation. PMID:27150753

  18. Systemic Metabolomic Changes in Blood Samples of Lung Cancer Patients Identified by Gas Chromatography Time-of-Flight Mass Spectrometry

    OpenAIRE

    Suzanne Miyamoto; Taylor, Sandra L.; Barupal, Dinesh K; Ayumu Taguchi; Gert Wohlgemuth; Wikoff, William R.; Yoneda, Ken Y.; Gandara, David R.; Samir M. Hanash; Kyoungmi Kim; Oliver Fiehn

    2015-01-01

    Lung cancer is a leading cause of cancer deaths worldwide. Metabolic alterations in tumor cells coupled with systemic indicators of the host response to tumor development have the potential to yield blood profiles with clinical utility for diagnosis and monitoring of treatment. We report results from two separate studies using gas chromatography time-of-flight mass spectrometry (GC-TOF MS) to profile metabolites in human blood samples that significantly differ from non-small cell lung cancer ...

  19. Engineering considerations for corrosion monitoring of gas gathering pipeline systems

    Energy Technology Data Exchange (ETDEWEB)

    Braga, T.G.; Asperger, R.G.

    1987-01-01

    Proper corrosion monitoring of gas gathering pipelines requires a system review to determine the appropriate monitor locations and types of monitoring techniques. This paper develops and discusses a classification of conditions such as flow regime and gas composition. Also discussed are junction categories which, for corrosion monitoring, need to be considered from two points of view. The first is related to fluid flow in the line and the second is related corrosion inhibitor movement along the pipeline. The appropriate application of the various monitoring techniques such as coupons, hydrogen detectors, electrical resistance probe and linear polarization probes are discussed in relation to flow regime and gas composition. Problems caused by semi-conduction from iron sulfide are considered. Advantages and disadvantages of fluid gathering methods such as pots and flow-through drips are discussed in relation to their reliability as on-line monitoring locations.

  20. Non-invasive respiratory monitoring in paediatric intensive care unit.

    OpenAIRE

    Nadkarni U; Shah A; Deshmukh C

    2000-01-01

    Monitoring respiratory function is important in a Paediatrics Intensive Care Unit (PICU), as majority of patients have cardio-respiratory problems. Non-invasive monitoring is convenient, accurate, and has minimal complications. Along with clinical monitoring, oxygen saturation using pulse oximetry, transcutaneous oxygenation (PtcO2) and transcutaneous PCO2 (PtcCO2) using transcutaneous monitors and end-tidal CO2 using capnography are important and routine measurements done in most PICUs. Cons...

  1. Coal mine gas monitoring system based on wireless sensor network

    Institute of Scientific and Technical Information of China (English)

    WANG Jian; WANG Ru-lin; WANG Xue-min; SHEN Chuan-he

    2007-01-01

    Based on the nowadays'condition.it is urgent that the gas detection cable communication system must be replaced by the wireless communication systems.The wireless sensors distributed in the environment can achieve the intelligent gas monitoring system.Apply with multilayer data fuse to design working tactics,and import the artificial neural networks to analyze detecting result.The wireless sensors system communicates with the controI center through the optical fiber cable.All the gas sensor nodes distributed in coal mine are combined into an intelligent,flexible structure wireless network system.forming coal mine gas monitoring system based on wireless sensor network.

  2. A multichannel bioimpedance monitor for full-body blood flow monitoring.

    Science.gov (United States)

    Vondra, Vlastimil; Jurak, Pavel; Viscor, Ivo; Halamek, Josef; Leinveber, Pavel; Matejkova, Magdalena; Soukup, Ladislav

    2016-02-01

    The design, properties, and possible diagnostic contribution of a multichannel bioimpedance monitor (MBM) with three independent current sources are presented in this paper. The simultaneous measurement of bioimpedance at 18 locations (the main part of the body, legs, arms, and neck) provides completely new information, on the basis of which more precise haemodynamic parameters can be obtained. The application of the MBM during various haemodynamic stages, such as resting in a supine position, tilting, exercise stress, and various respiration manoeuvres, is demonstrated. Statistical analysis on a group of 34 healthy volunteers is presented for demonstration of blood flow monitoring by using the proposed method. PMID:25992508

  3. Monitoring of rapid blood pH variations by CO detection in breath with tunable diode laser

    Science.gov (United States)

    Kouznetsov, Andrian I.; Stepanov, Eugene V.; Zyrianov, Pavel V.; Shulagin, Yurii A.; Diachenko, Alexander I.; Gurfinkel, Youri I.

    1997-06-01

    Detection of endogenous carbon monoxide content in breath with tunable diode lasers (TDL) was proposed for noninvasive monitoring of rapid blood pH variation. Applied approach is based on high sensitivity of the haemoglobin and myoglobin affinity for CO to blood pH value and an ability to detect rapidly small variations of CO content in expired air. Breath CO absorption in 4.7 micrometers spectral region was carefully measured using PbSSe tunable diode laser that can provide 1 ppb CO concentration sensitivity and 10 s time constant. Applied TDL gas analyzer was used to monitor expired air of studied persons in physiological tests including hyperventilation and physical load. Simultaneous blood tests were conducted to demonstrate correlation between blood and breath chemical parameters.

  4. Landfill gas monitoring at borehole wells using an autonomous environmental monitoring system

    OpenAIRE

    Kiernan, Breda M.; Beirne, Stephen; Fay, Cormac; Diamond, Dermot

    2008-01-01

    An autonomous environmental monitoring system(Smart Landfill) has been constructed for the quantitative measurement of the components of landfill gas found at borehole wells at the perimeter of landfill sites. The main components of landfill gas are the greenhouse gases, methane and carbon dioxide and have been monitored in the range 0-5 % volume. This monitoring system has not only been tested in the laboratory but has been deployed in multiple field trials and the data collected successf...

  5. Hybrid CARS for Non-Invasive Blood Glucose Monitoring

    Science.gov (United States)

    Wang, Xi; Pestov, Dmitry; Zhang, Aihua; Murawski, Robert; Sokolov, Alexei; Welch, George; Laane, Jaan; Scully, Marlan

    2007-10-01

    We develop a spectroscopy technique that combines the advantages of both the frequency-resolved coherent anti-Stokes Raman scattering (CARS) and the time-resolved CARS. We use broadband preparation pulses to get an instantaneous coherent excitation of multiplex molecular vibration levels and subsequent optically shaped time-delayed narrowband probing pulse to detect these vibrations. This technique can suppress the nonresonant background and retrieve the molecular fingerprint signal efficiently and rapidly. We employ this technique to glucose detection, the final goal of which is accurate, non-invasive (i.e. painless) and continuous monitoring of blood glucose concentration in the Diabetes diagnosis to replace the current glucose measurement process, which requires painful fingerpricks and therefore cannot be performed more than a few times a day. We have gotten the CARS spectra of glucose aqueous solution down to 2 mM.

  6. Blood Pressure Mobile Monitoring for Pregnant Woman Based Android System

    Science.gov (United States)

    Supriyanti, Retno; Erfayanto, Uji; Ramadani, Yogi; Murdyantoro, Eko; Widodo, Haris B.

    2016-01-01

    Currently, at least 18,000 women die every year in Indonesia due to pregnancy or childbirth. It means that every half hour a woman dies due to pregnancy or childbirth. As a result, every year 36,000 children became orphans. The high maternal mortality rate was put Indonesia on top in ASEAN. The main causes of maternal mortality are high-risk pregnancy. Mothers who have diseases like high blood pressure, pre-eclampsia, diabetes, hyperthyroidism, and already over 40 years old and infectious diseases such as rubella, hepatitis and HIV can be factors that lead to high-risk pregnancy. This paper will discuss the development of a blood pressure monitoring device that is suitable for pregnant women. It is based on convenience for pregnant women to get the equipment that is flexible with her presence. Results indicate that the equipment is in use daily support for pregnant women therefore, one of the causes of maternal mortality can be detected earlier.

  7. Ambulatory blood pressure monitoring in solid organ transplantation.

    Science.gov (United States)

    Ramesh Prasad, G V

    2012-01-01

    Solid organ transplant recipients are at an increased risk for hypertension and cardiovascular disease. To assist in their management, 24-h ambulatory blood pressure monitoring (ABPM) has become increasingly used in both clinical research settings and practice. ABPM has been used to better define post-transplant hypertension incidence and prevalence in different solid organ transplantation populations. ABPM provides additional information on cardiovascular risk beyond that obtained by clinic-based readings, based on its ability to assess 24-h blood pressure (BP) load, detect nocturnal non-dipping, and predict target organ damage. It has provided some assurance about the safety of living kidney donation. Information from ABPM can be used to guide living kidney donor selection, and because ABPM-related data has been correlated with clinically important kidney and heart transplant recipient outcomes, it may be a valuable adjunct in their management. Despite these advantages, barriers to wider use of ABPM include expense, clinical inertia in hypertension management, lack of prospective clinical trial data, and clinical problems that compete with hypertension for attention such as acute or chronic allograft dysfunction. The increasing amount of research and clinical use for ABPM may allow for closer assessment and intervention to help address the increased cardiovascular risk faced by many solid organ transplant recipients. PMID:22220828

  8. Gas pipeline monitoring in Europe by Satellite SAR

    NARCIS (Netherlands)

    Smith, A.J.E.

    2002-01-01

    At present, gas pipeline networks in Europe are routinely monitored by vehicle and air patrols to protect them against damage by soil movement and third part interference. Because of the expenses, pipeline operators are investigating the possibilities to replace these traditional monitoring methods

  9. Mobile Personal Health System for Ambulatory Blood Pressure Monitoring

    Directory of Open Access Journals (Sweden)

    Luis J. Mena

    2013-01-01

    Full Text Available The ARVmobile v1.0 is a multiplatform mobile personal health monitor (PHM application for ambulatory blood pressure (ABP monitoring that has the potential to aid in the acquisition and analysis of detailed profile of ABP and heart rate (HR, improve the early detection and intervention of hypertension, and detect potential abnormal BP and HR levels for timely medical feedback. The PHM system consisted of ABP sensor to detect BP and HR signals and smartphone as receiver to collect the transmitted digital data and process them to provide immediate personalized information to the user. Android and Blackberry platforms were developed to detect and alert of potential abnormal values, offer friendly graphical user interface for elderly people, and provide feedback to professional healthcare providers via e-mail. ABP data were obtained from twenty-one healthy individuals (>51 years to test the utility of the PHM application. The ARVmobile v1.0 was able to reliably receive and process the ABP readings from the volunteers. The preliminary results demonstrate that the ARVmobile 1.0 application could be used to perform a detailed profile of ABP and HR in an ordinary daily life environment, bedsides of estimating potential diagnostic thresholds of abnormal BP variability measured as average real variability.

  10. Development and Remodeling of the Vertebrate Blood-Gas Barrier

    Directory of Open Access Journals (Sweden)

    Andrew Makanya

    2013-01-01

    Full Text Available During vertebrate development, the lung inaugurates as an endodermal bud from the primitive foregut. Dichotomous subdivision of the bud results in arborizing airways that form the prospective gas exchanging chambers, where a thin blood-gas barrier (BGB is established. In the mammalian lung, this proceeds through conversion of type II cells to type I cells, thinning, and elongation of the cells as well as extrusion of the lamellar bodies. Subsequent diminution of interstitial tissue and apposition of capillaries to the alveolar epithelium establish a thin BGB. In the noncompliant avian lung, attenuation proceeds through cell-cutting processes that result in remarkable thinning of the epithelial layer. A host of morphoregulatory molecules, including transcription factors such as Nkx2.1, GATA, HNF-3, and WNT5a; signaling molecules including FGF, BMP-4, Shh, and TFG-β and extracellular proteins and their receptors have been implicated. During normal physiological function, the BGB may be remodeled in response to alterations in transmural pressures in both blood capillaries and airspaces. Such changes are mitigated through rapid expression of the relevant genes for extracellular matrix proteins and growth factors. While an appreciable amount of information regarding molecular control has been documented in the mammalian lung, very little is available on the avian lung.

  11. Operating experience review of an INL gas monitoring system

    Energy Technology Data Exchange (ETDEWEB)

    Cadwallader, Lee C. [Idaho National Lab. (INL), Idaho Falls, ID (United States); DeWall, K. G. [Idaho National Lab. (INL), Idaho Falls, ID (United States); Herring, J. S. [Idaho National Lab. (INL), Idaho Falls, ID (United States)

    2015-03-12

    This article describes the operations of several types of gas monitors in use at the Idaho National Laboratory (INL) High Temperature Electrolysis Experiment (HTE) laboratory. The gases monitored in the lab room are hydrogen, carbon monoxide, carbon dioxide, and oxygen. The operating time, calibration, and both actual and unwanted alarms are described. The calibration session time durations are described. In addition, some simple calculations are given to estimate the reliability of these monitors and the results are compared to operating experiences of other types of monitors.

  12. Analysis of mean transcutaneous capnography in consecutive patients undergoing polysomnography

    OpenAIRE

    Giulio Cesare Pinnola; Patrícia Souza Bastos

    2014-01-01

    Transcutaneous capnography is a noninvasive method useful for analysis of the behavioral tendency of transcutaneous CO2 pressure (PtcCO2) in patients undergoing polysomnography, to evaluate respiratory sleep disorders. Objective Determine normative PtcCO2 values in normal patients undergoing polysomnography. Method One hundred seventy-nine patients who underwent polysomnography with simultaneous PtcCO2 measurement were assessed by means of a transcutaneous capnograph (TCM4 series from Radi...

  13. Monitoring of noble gas radioisotopes in nuclear power plant effluents

    International Nuclear Information System (INIS)

    Monitoring of gaseous radionuclides in the effluents of nuclear facilities is an essential requirement in effluent management programs. Since there is no practical way of removing noble gas radioisotopes from air at release pathways, their accurate monitoring is essential for providing appropriate environmental protection. Emitted γ dose-rate is the limiting factor for concentration-time integral of noble gas in gaseous effluents of reactor facilities. The external exposure to the public from a semi-infinite cloud is directly proportional to both the noble gas isotope concentration and the integrated γ energy per disintegration. Both can be directly measured in gaseous effluent pathways with a suitable detector. The capability of NaI(T1), CaF2(Eu) and plastic scintillation detectors to measure the γ-Ci.MeV content of noble gas releases was experimentally evaluated. The combination of CaF2(Eu) detector in a pressurized through-flow chamber, with a charge integrating scaler well complied with both γ energy response and detection sensitivity requirements. Noble gas source terms and effluent monitoring criteria are discussed, theoretical and experimental results are presented and a practical, on-line noble gas monitoring system is described

  14. Development of oil and gas sector monitoring in Perm territory

    Directory of Open Access Journals (Sweden)

    Galina Vasil'evna Kutergina

    2012-03-01

    Full Text Available This paper reviews current approaches to the definition of «regional monitoring» and its contents. The work is based on the use of a systematic approach to the analysis of regional monitoring, reviewing it as part of the overall control system and risk management in the region.Organization of regional monitoring is considered on the example of oil and gas complex (OGC of Perm territory. This paper summarizes the structure of the OGC, the specific features of the activities of the enterprises that have the most significant impact on the organization of monitoring. The findings are based on an analysis and compilation of statistics. Authors consider in most details the subjects and objects of state and corporate level monitoring of the regional OGC in Perm territory, their main function of monitoring, interoperability issues, methodological support of various institutions in the periodic monitoring of OGC - the audit committees and internal audit units. Proposals for the development in most parts refer to the use of risk-oriented approach to organizing periodic monitoring of oil and gas industry in the territory on the basis of a common methodology for assessing its effectiveness. The proposals to expand cooperation between state agencies and regional bodies of governance of OGC enterprises in Perm territory on a wide range of areas of the organization of monitoring: the exchange of professional information, methodology, activities, staff and others.

  15. Ambulatory Blood Pressure Monitoring-Derived Short-Term Blood Pressure Variability in Primary Aldosteronism.

    Science.gov (United States)

    Grillo, Andrea; Bernardi, Stella; Rebellato, Andrea; Fabris, Bruno; Bardelli, Moreno; Burrello, Jacopo; Rabbia, Franco; Veglio, Franco; Fallo, Francesco; Carretta, Renzo

    2015-08-01

    The aim of this study was to investigate the short-term blood pressure (BP) variability (BPV) derived from ambulatory blood pressure monitoring (ABPM) in patients with primary aldosteronism (PA), either idiopathic hyperaldosteronism (IHA) or aldosterone-producing adenoma (APA), in comparison with patients with essential hypertension (EH) and normotensive (NT) controls. Thirty patients with PA (16 with IHA and 14 with APA), 30 patients with EH, and 30 NT controls, matched for sex, age, body mass index, and antihypertensive therapy, were studied. The standard deviation (SD) of 24-hour, daytime, and nighttime BP; 24-hour weighted SD of BP; and 24-hour BP average real variability were not different between patients with PA and those with EH (P=not significant). All BPV indices were higher in patients with PA, either IHA or APA subtypes, and patients with EH, compared with NT controls (P<.001 to P<.05). ABPM-derived short-term BPV is increased in patients with PA, and it may represent an additional cardiovascular risk factor in this disease. The role of aldosterone excess in BPV has to be clarified. PMID:25880017

  16. A Gas Monitoring and Control System in a Coal and Gas Outburst Laboratory

    Directory of Open Access Journals (Sweden)

    W. Nie

    2014-01-01

    Full Text Available Coal and gas outburst is a phenomenon characterized by the ejection of gas and coal from the solid face of a mine. Physical minioutburst experiments are a very important tool for analyzing outbursts of coal and gas. However, few reports have focused on the safety problem produced by gas concentration or the role of gas spread during the physical experiments. In this study, we designed a simple monitoring and control system for the safety of staff during the minioutburst experiments. The results showed that, in the simulation of four dangerous situations, the system based on a sensors feedback loop can monitor the development of gas content in the temporal and spatial domains for the enhancement of accurate warnings. The system also automatically chooses the appropriate ventilation measures to lower the gas content considering different degrees of danger.

  17. Evaluation of the Capillary Blood Glucose Self-monitoring Program

    Directory of Open Access Journals (Sweden)

    Mariana Cristina Augusto

    2014-10-01

    Full Text Available OBJECTIVE: to evaluate the structure, process and results of the Capillary Blood Glucose Self-monitoring Program in a Brazilian city.METHOD: epidemiological, cross-sectional study. The methodological framework of Donabedian was used to construct indicators of structure, process and outcome. A random sample (n = 288 of users enrolled and 96 health professionals who worked in the program was studied. Two questionnaires were used that were constructed for this study, one for professionals and one for users, both containing data for the evaluation of structure, process and outcome. Anthropometric measures and laboratory results were collected by consulting the patients' health records. The analysis involved descriptive statistics.RESULTS: most of the professionals were not qualified to work in the program and were not knowledgeable about the set of criteria for patient registration. None of the patients received complete and correct orientations about the program and the percentage with skills to perform conducts autonomously was 10%. As regards the result indicators, 86.4% of the patients and 81.3% of the professionals evaluated the program positively.CONCLUSION: the evaluation indicators designed revealed that one of the main objectives of the program, self-care skills, has not been achieved.

  18. Mucosal blood flow measurements using laser Doppler perfusion monitoring

    Institute of Scientific and Technical Information of China (English)

    Dag Arne Lihaug Hoff; Hans Gregersen; Jan Gunnar Hatlebakk

    2009-01-01

    Perfusion of individual tissues is a basic physiological process that is necessary to sustain oxygenation and nutrition at a cellular level. Ischemia, or the insufficiency of perfusion, is a common mechanism for tissue death or degeneration, and at a lower threshold, a mechanism for the generation of sensory signalling including pain. It is of considerable interest to study perfusion of peripheral abdominal tissues in a variety of circumstances. Microvascular disease of the abdominal organs has been implicated in the pathogenesis of a variety of disorders, including peptic ulcer disease, inflammatory bowel disease and chest pain. The basic principle of laser Doppler perfusion monitoring (LDPM) is to analyze changes in the spectrum of light reflected from tissues as a response to a beam of monochromatic laser light emitted. It reflects the total local microcirculatory blood perfusion, including perfusion in capillaries, arterioles, venules and shunts. During the last 20-25 years, numerous studies have been performed in different parts of the gastrointestinal (GI) tract using LDPM. In recent years we have developed a multi-modal catheter device which includes a laser Doppler probe, with the intent primarily to investigate patients suffering from functional chest pain of presumed oesophageal origin. Preliminary studies show the feasibility of incorporating LDPM into such catheters for performing physiological studies in the GI tract. LDPM has emerged as a research and clinical tool in preference to other methods; but, it is important to be aware of its limitations and account for them when reporting results.

  19. Dried blood spot analysis for therapeutic drug monitoring of pazopanib.

    Science.gov (United States)

    de Wit, Djoeke; den Hartigh, Jan; Gelderblom, Hans; Qian, Yanwen; den Hollander, Margret; Verheul, Henk; Guchelaar, Henk-Jan; van Erp, Nielka P

    2015-12-01

    Dried blood spot (DBS) sampling is potentially a more patient-friendly and flexible alternative to venous sampling of pazopanib. This study determined the agreement between pazopanib DBS and plasma concentrations to facilitate implementation of pazopanib DBS sampling into clinical practice. Paired DBS and plasma samples were collected in 12 patients. Pazopanib plasma concentrations were calculated from DBS concentrations using the formula: plasma concentration = DBSconcentration /(1 - hematocrit). Passing-Bablok and Bland-Altman analyses were used to determine the agreement between calculated and measured plasma concentrations. We predefined a clinical acceptance limit of 25% for the Bland-Altman analysis. Passing-Bablok analysis showed a small constant (intercept estimate, -8.53 [95%CI, -12.22 to -4.41]) and slightly proportional (slope estimate, 1.15 [95%CI, 1.04-1.24]) bias between calculated and measured concentrations. This bias was clinically nonrelevant, as shown by Bland-Altman analysis; the mean ratio of calculated to measured concentrations was 0.94 (95%CI, 0.65-1.23). The clinical acceptance limits were well within these 95% limits of agreement. More specifically, 92.6% of the data points were within the predefined acceptance limits. Pazopanib plasma concentrations can be accurately calculated from DBS concentrations. Although validation of DBS cards prepared by patients themselves is required, these results show that DBS sampling can be used to monitor pazopanib therapy in clinical practice. PMID:26032288

  20. Infrared spectroscopy for monitoring gas hydrates in aqueous solution

    Energy Technology Data Exchange (ETDEWEB)

    Dobbs, G.T.; Luzinova, Y.; Mizaikoff, B. [Georgia Inst. of Technology, Atlanta, GA (United States). School of Chemistry and Biochemistry; Raichlin, Y.; Katzir, A. [Tel-Aviv Univ., Tel-Aviv (Israel). Shool of Physics and Astronomy

    2008-07-01

    This paper introduced the first principles for monitoring gas hydrate formation and dissociation in aqueous solution by evaluating state-responsive infrared (IR) absorption features of water with fiberoptic evanescent field spectroscopy. A first order linear functional relationship was also derived according to Lambert Beer's law in order to quantify the percentage gas hydrate within the volume of water probed via the evanescent field. In addition, spectroscopic studies evaluating seafloor sediments collected from a gas hydrate site in the Gulf of Mexico revealed minimal spectral interferences from sediment matrix components. As such, evanescent field sensing strategies were established as a promising perspective for monitoring the dynamics of gas hydrates in oceanic environments. 21 refs., 5 figs.

  1. The influence of acupoint transcutaneous electrical stimulation on blood glucose, glycosylated hemoglobin and the physical fitness of patients with type 2 diabetes mellitus%经皮穴位电刺激疗法对2型糖尿病患者血糖和糖化血红蛋白及体质指标的影响

    Institute of Scientific and Technical Information of China (English)

    杨铭; 吴志远; 贾杰; 吴毅; 黄天生; 苏红梅; 王松伟; 陆佳敏; 吕颖

    2014-01-01

    Objective To compare the effects of electrical stimulation applied transcutaneously to acupoints with electroacupuncture on blood glucose,glycosylated hemoglobin levels and the physical fitness of type 2 diabetic patients.Methods Sixty persons with type 2 diabetes were randomized into an electroacupuncture group (30 cases) and a transcutaneous stimulation group (30 cases).The patients were treated with either electroacupuncture or transcutaneous electrical stimulation of certain acupoints for a period of 2 months,30 minutes a day,5 times per week.At pre-treatment,post-treatment and follow-up,the indexes of 2 hour postprandial blood glucose (2hPG) and glycosylated hemoglobin (HbA1c),body mass index (BMI),waist-to-hip ratio (WHR) and percentage of body fat (PBF) were measured.Results After treatment,all of the above indexes had improved in both groups.At follow-up there was no longer any statistically significant difference compared with before treatment except for HbA1c levels in the electroacupuncture group.There was no statistically significant difference between the groups in terms of any index at any time point.Conclusion Transcutaneous electrical stimulation of the acupoints demonstrated effects similar to those of electroacupuncture in improving the blood glucose and glycosylated hemoglobin levels and the physical fitness of patients with type 2 diabetes.%目的 探讨经皮穴位电刺激疗法和电针疗法对2型糖尿病患者血糖、糖化血红蛋白及体质指标的影响.方法 将60例2型糖尿病患者按随机数字表法分为电针组和Acu-TENS组,每组30例.2组患者在常规用药基础上,分别进行为期2个月,每日1次,每次30 min,每周5次的电针治疗和经皮穴位电刺激治疗.分别于治疗前、治疗2个月后(治疗后)及治疗结束后2个月(随访时)对2组患者进行餐后2h血糖、糖化血红蛋白、体质指数(BMI)、腰臀比、体脂百分比等指标检测,并进行统计学分析比较.结果 治

  2. Real-time trend monitoring of gas compressor stations

    Energy Technology Data Exchange (ETDEWEB)

    Van Hardeveld, T. (Nova, an Alberta Corp., AB (Canada))

    1991-02-01

    The authors' company has developed a machinery health monitoring system (MHealth) for short-term and long-term historical trending and analysis of data from its 40 gas compressor stations. The author discusses the benefits of real-time trending in troubleshooting operations, in preventative maintenance scheduling and cites specific applications in the startup operations of several new gas compressor/centrifugal compressor units.

  3. Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia

    OpenAIRE

    Meng, Xianghu; ZANG, GUANGHUI; FAN, LONGCHANG; Zheng, Lei; DAI, JINZHEN; WANG, XUEREN; Xia, Wei; Liu, Jihong; ZHANG, CHUANHAN

    2013-01-01

    The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscil...

  4. Development of an autonomous greenhouse gas monitoring system

    OpenAIRE

    Kiernan, Breda M.; Fay, Cormac; Beirne, Stephen; Diamond, Dermot

    2008-01-01

    This paper describes the designs of a first and second generation autonomous gas monitoring system and the successful field trial of the final system (2nd generation). Infrared sensing technology is used to detect and measure the greenhouse gases methane (CH4) and carbon dioxide (CO2) at point sources. The ability to monitor real-time events is further enhanced through the implementation of both GSM and Bluetooth technologies to communicate these data in real-time. These systems are robust...

  5. Transcutaneous absorption of oil in preterm babies--a pilot study.

    Science.gov (United States)

    Fernandez, A R; Krishnamoorthy, G; Patil, N; Mondkar, J A; Swar, B D

    2005-03-01

    This study was conducted to determine transcutaneous absorption of oil in preterm neonates. A mixture of coconut oil and Meadowfoam oil which contains unique fatty acids, which acted as marker fatty acids was applied to the skin of babies. One mL blood was collected before and one hour after post oil application. Both pre and post oil application serum samples were hydrolysed and derivatised with 2-phenyl hydrazine hydrochloride in order to detect fatty acids by HPLC analysis on C-8 column. None of the pre oil application serum sample showed the presence of the marker fatty acids. The post oil application serum sample of all the 12 babies showed the presence of marker fatty acids of Meadowfoam oil which indicates transcutaneous absorption of oil in preterm babies.

  6. Results of ambulatory arterial blood pressure monitoring in children with obesity

    OpenAIRE

    Faruk Öktem

    2010-01-01

    Objectives: The relationship between obesity and essential hypertension is well known. In this study, we aimed to evaluate ambulatory arterial blood pressure monitoring of obese and non-obese children who had similar demographic characteristics.Materials and methods: Seventy one children and adolescents (n=39 obesity, n=32 controls) were studied. Blood pressure of the children were measured by 24 hour ambulatory blood pressure monitoring device.Results: Obese children had significantly higher...

  7. Software to Control and Monitor Gas Streams

    Science.gov (United States)

    Arkin, C.; Curley, Charles; Gore, Eric; Floyd, David; Lucas, Damion

    2012-01-01

    This software package interfaces with various gas stream devices such as pressure transducers, flow meters, flow controllers, valves, and analyzers such as a mass spectrometer. The software provides excellent user interfacing with various windows that provide time-domain graphs, valve state buttons, priority- colored messages, and warning icons. The user can configure the software to save as much or as little data as needed to a comma-delimited file. The software also includes an intuitive scripting language for automated processing. The configuration allows for the assignment of measured values or calibration so that raw signals can be viewed as usable pressures, flows, or concentrations in real time. The software is based on those used in two safety systems for shuttle processing and one volcanic gas analysis system. Mass analyzers typically have very unique applications and vary from job to job. As such, software available on the market is usually inadequate or targeted on a specific application (such as EPA methods). The goal was to develop powerful software that could be used with prototype systems. The key problem was to generalize the software to be easily and quickly reconfigurable. At Kennedy Space Center (KSC), the prior art consists of two primary methods. The first method was to utilize Lab- VIEW and a commercial data acquisition system. This method required rewriting code for each different application and only provided raw data. To obtain data in engineering units, manual calculations were required. The second method was to utilize one of the embedded computer systems developed for another system. This second method had the benefit of providing data in engineering units, but was limited in the number of control parameters.

  8. Effect of administration of water enriched in O2 by injection or electrolysis on transcutaneous oxygen pressure in anesthetized pigs

    Directory of Open Access Journals (Sweden)

    Charton A

    2014-08-01

    Full Text Available Antoine Charton,1 François Péronnet,2 Stephane Doutreleau,3 Evelyne Lonsdorfer,3 Alexis Klein,4 Liliana Jimenez,4 Bernard Geny,3 Pierre Diemunsch,1 Ruddy Richard5 1Department of Anesthesia and Critical Care, and EA 3072, Hôpital de Hautepierre; University of Strasbourg, Strasbourg, France; 2Department of Kinesiology, Université de Montréal, Montreal, QC, Canada; 3CHRU of Strasbourg, Physiology and Functional Explorations Department, New Civil Hospital, Strasbourg, France and University of Strasbourg, Faculty of Medicine, Physiology Department, Strasbourg, France; 4Danone Research, Palaiseau, France; 5Department of Sport Medicine and Functional Explorations, CHU Clermont-Ferrand and INRA UMR 1019, CRNH-Auvergne, Clermont-Ferrand, France Background: Oral administration of oxygenated water has been shown to improve blood oxygenation and could be an alternate way for oxygen (O2 supply. In this experiment, tissue oxygenation was compared in anesthetized pigs receiving a placebo or water enriched in O2 by injection or a new electrolytic process. Methods: Forty-two pigs randomized in three groups received either mineral water as placebo or water enriched in O2 by injection or the electrolytic process (10 mL/kg in the stomach. Hemodynamic parameters, partial pressure of oxygen in the arterial blood (PaO2, skin blood flow, and tissue oxygenation (transcutaneous oxygen pressure, or TcPO2 were monitored during 90 minutes of general anesthesia. Absorption and tissue distribution of the three waters administered were assessed using dilution of deuterium oxide. Results: Mean arterial pressure, heart rate, PaO2, arteriovenous oxygen difference, and water absorption from the gut were not significantly different among the three groups. The deuterium to protium ratio was also similar in the plasma, skin, and muscle at the end of the protocol. Skin blood flow decreased in the three groups. TcPO2 slowly decreased over the last 60 minutes of the experiment in

  9. Analysis of blood gas values in mice following pulmonary irradiation

    International Nuclear Information System (INIS)

    The arterial pH and partial pressures of oxygen (P/sub a/O2) and carbon dioxide (P/sub a/CO2) of mice were investigated at times from 1 to 16 weeks after a single radiation dose of 1100 rad to the thorax. The mice breathed air during the irradiation and either air, 7% O2:93% N2, or 100% O2 during the blood gas determinations. P/sub a/O2 values of mice breathing 7% O2:93% N2 or 100% O2 were reduced 4 to 6 weeks after the irradiation. Also, all P/sub a/O2 values were reduced and all hemoglobin (Hb) levels were elevated 16 weeks after irradiation. The majority of the animals died between 16 and 26 weeks after irradiation, but decreased P/sub a/O2 and increased Hb levels also were observed 27 weeks after the treatment in a few of the surviving mice. These findings indicate that the arterial blood gases may change following pulmonary irradiation and may provide an endpoint for the study of radiation-induced pulmonary damage

  10. Chronobiologically Interpreted Ambulatory Blood Pressure Monitoring in Health and Disease.

    Science.gov (United States)

    Halberg, Franz; Mult, Hc; Cornélissen, Germaine; Hillman, Dewayne; Beaty, Larry A; Hong, Shiyu; Schwartzkopff, Othild; Watanabe, Yoshihiko; Otsuka, Kuniaki; Siegelova, Jarmila

    2012-05-01

    To detect vascular variability anomalies (VVAs), a blood pressure and heart rate profile around the clock for at least 7 days is a start. As a minimum, measurement every 60 or preferably 30 minutes for a week is needed, to be continued if abnormality is found, to assess the about 24-hour (circadian) variability that exists in all individuals. As a first dividend, one then also obtains a glimpse of 2 of the very many longer-than-circadian periodicities, the biological half-week and week. Certainly if we can have sensors and computer chips in our cars that continuously monitor the pressure over a tire's life, we should be able to do the same job for ourselves for diagnostic and therapeutic decisions. Healthcare today emphasizes wellness with recommendations for exercise and a proper diet, yet these evaluations may not be adequate. BP may be measured at a visit to the doctor or before an exercise session, along with measuring body weight and performing a physical exam. The seeds of disease are planted long before they are visible, and what appears to be normal from a conventional point of view may in fact actually be abnormal. Hidden alterations of physiological function, masked by the body's remarkable adaptive capabilities, may become visible through a new diagnostic and therapeutic realm-chronobiology-that reveals hitherto unseen abnormalities. The tools of chronobiology may yield additional dividends, such as the detection of physiological "loads" related to stress and stress relief and the undesirable effcts of space weather upon personal events such as sudden cardiac death, societal events like terrorism and war, and natural disasters. Chronobiologi cally interpreted automatic ambulatory BP and heart rate (HR) monitoring (C-ABPM) may detect the antecedents of these types of events. C-ABPM is of interest in preventive cardiology, since it reveals new diagnoses as vascular variability anomalies (VVAs) and renders previous conventional diagnoses more reliable, such

  11. TRANSCUTANEOUS ELECTRICAL NERVE-STIMULATION (TENS) IN RAYNAUDS-PHENOMENON

    NARCIS (Netherlands)

    MULDER, P; DOMPELING, EC; VANSLOCHTERENVANDERBOOR, JC; KUIPERS, WD; SMIT, AJ

    1991-01-01

    Transcutaneous nerve stimulation (TENS) has been described as resulting in vasodilatation. The effect of 2 Hz TENS of the right hand during forty-five minutes on skin temperature and plethysmography of the third digit of both hands and feet and on transcutaneous oxygen tension (TcpO2) of the right h

  12. Wearable Beat to Beat Blood Pressure Monitor Project

    Data.gov (United States)

    National Aeronautics and Space Administration — A key component of NASA's human exploration programs is a system that monitors the health of the crew during the space missions. The wearable physiological monitor...

  13. The effect of transcutaneous application of carbon dioxide (CO{sub 2}) on skeletal muscle

    Energy Technology Data Exchange (ETDEWEB)

    Oe, Keisuke [Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, Kobe (Japan); Ueha, Takeshi [NeoChemir Inc, Kobe (Japan); Sakai, Yoshitada, E-mail: sakai.yoshitada@gm.himeji-du.ac.jp [Faculty of Health Care Sciences, Himeji Dokkyo University, Himeji (Japan); Niikura, Takahiro; Lee, Sang Yang; Koh, Akihiro [Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, Kobe (Japan); Hasegawa, Takumi [Department of Oral and Maxillofacial Surgery, Kobe University Graduate School of Medicine, Kobe (Japan); Tanaka, Masaya [NeoChemir Inc, Kobe (Japan); Miwa, Masahiko; Kurosaka, Masahiro [Department of Orthopaedic Surgery, Kobe University Graduate School of Medicine, Kobe (Japan)

    2011-04-01

    Highlights: {yields} PGC-1{alpha} is up-regulated as a result of exercise such as mitochondrial biogenesis and muscle fiber-type switching, and up-regulation of VEGF. {yields} We demonstrated transcutaneous application of CO{sub 2} up-regulated the gene expression of PGC-1{alpha}, SIRT1 and VEGF, and instance of muscle fiber switching. {yields} Transcutaneous application of CO{sub 2} may cause similar effect to aerobic exercise in skeletal muscle. -- Abstract: In Europe, carbon dioxide therapy has been used for cardiac disease and skin problems for a long time. However there have been few reports investigating the effects of carbon dioxide therapy on skeletal muscle. Peroxisome proliferators-activated receptor (PPAR)-gamma coactivator-1 (PGC-1{alpha}) is up-regulated as a result of exercise and mediates known responses to exercise, such as mitochondrial biogenesis and muscle fiber-type switching, and neovascularization via up-regulation of vascular endothelial growth factor (VEGF). It is also known that silent mating type information regulation 2 homologs 1 (SIRT1) enhances PGC-1{alpha}-mediated muscle fiber-type switching. Previously, we demonstrated transcutaneous application of CO{sub 2} increased blood flow and a partial increase of O{sub 2} pressure in the local tissue known as the Bohr effect. In this study, we transcutaneously applied CO{sub 2} to the lower limbs of rats, and investigated the effect on the fast muscle, tibialis anterior (TA) muscle. The transcutaneous CO{sub 2} application caused: (1) the gene expression of PGC-1{alpha}, silent mating type information regulation 2 homologs 1 (SIRT1) and VEGF, and increased the number of mitochondria, as proven by real-time PCR and immunohistochemistry, (2) muscle fiber switching in the TA muscle, as proven by isolation of myosin heavy chain and ATPase staining. Our results suggest the transcutaneous application of CO{sub 2} may have therapeutic potential for muscular strength recovery resulting from disuse

  14. Remote Real-Time Monitoring of Subsurface Landfill Gas Migration

    Directory of Open Access Journals (Sweden)

    Alan F. Smeaton

    2011-06-01

    Full Text Available The cost of monitoring greenhouse gas emissions from landfill sites is of major concern for regulatory authorities. The current monitoring procedure is recognised as labour intensive, requiring agency inspectors to physically travel to perimeter borehole wells in rough terrain and manually measure gas concentration levels with expensive hand-held instrumentation. In this article we present a cost-effective and efficient system for remotely monitoring landfill subsurface migration of methane and carbon dioxide concentration levels. Based purely on an autonomous sensing architecture, the proposed sensing platform was capable of performing complex analytical measurements in situ and successfully communicating the data remotely to a cloud database. A web tool was developed to present the sensed data to relevant stakeholders. We report our experiences in deploying such an approach in the field over a period of approximately 16 months.

  15. Remote Real-Time Monitoring of Subsurface Landfill Gas Migration

    Science.gov (United States)

    Fay, Cormac; Doherty, Aiden R.; Beirne, Stephen; Collins, Fiachra; Foley, Colum; Healy, John; Kiernan, Breda M.; Lee, Hyowon; Maher, Damien; Orpen, Dylan; Phelan, Thomas; Qiu, Zhengwei; Zhang, Kirk; Gurrin, Cathal; Corcoran, Brian; O’Connor, Noel E.; Smeaton, Alan F.; Diamond, Dermot

    2011-01-01

    The cost of monitoring greenhouse gas emissions from landfill sites is of major concern for regulatory authorities. The current monitoring procedure is recognised as labour intensive, requiring agency inspectors to physically travel to perimeter borehole wells in rough terrain and manually measure gas concentration levels with expensive hand-held instrumentation. In this article we present a cost-effective and efficient system for remotely monitoring landfill subsurface migration of methane and carbon dioxide concentration levels. Based purely on an autonomous sensing architecture, the proposed sensing platform was capable of performing complex analytical measurements in situ and successfully communicating the data remotely to a cloud database. A web tool was developed to present the sensed data to relevant stakeholders. We report our experiences in deploying such an approach in the field over a period of approximately 16 months. PMID:22163975

  16. Monitoring carbon dioxide in mechanically ventilated patients during hyperbaric treatment

    DEFF Research Database (Denmark)

    Bjerregård, Asger; Jansen, Erik

    2012-01-01

    Measurement of the arterial carbon dioxide (P(a)CO(2)) is an established part of the monitoring of mechanically ventilated patients. Other ways to get information about carbon dioxide in the patient are measurement of end-tidal carbon dioxide (P(ET)CO(2)) and transcutaneous carbon dioxide (PTCCO2......). Carbon dioxide in the blood and cerebral tissue has great influence on vasoactivity and thereby blood volume of the brain. We have found no studies on the correlation between P(ET)CO(2) or P(TC)CO(2), and P(a)CO(2) during hyperbaric oxygen therapy (HBOT)....

  17. The Food and Drug Administration Is Now Preparing to Establish Tighter Performance Requirements for Blood Glucose Monitors

    OpenAIRE

    Klonoff, David C.

    2010-01-01

    On March 16 and 17, 2010, the Food and Drug Administration (FDA) presented a public meeting about blood glucose monitoring at the Gaithersberg Hilton Hotel. The meeting was intended to present expert opinions and solicit input from the public about whether to develop new regulatory policies for blood glucose monitors. The meeting was divided into three sections: (1) Clinical Accuracy Requirements for Blood Glucose Monitors, (2) Interferences and Limitations of Blood Glucose Monitors, and (3) ...

  18. An elapsed time-temperature monitor for blood storage.

    Science.gov (United States)

    Harris, G E; Cloud, S; Myhre, B A

    1977-01-01

    Blood should not be allowed to exceed 10 C while being stored or transported. However, one cannot test the internal temperature of a unit of blood without contaminating it. Most blood banks have established an arbitrary time limit beyond which a blood unit cannot be kept out of the refrigerator. This method is ineffective if blood is stored in a satellite refrigerator, since the blood may be moved in and out of the refrigerator and the blood bank personnel will be unaware of it. An elapsed time indicator is described which employs a small condenser (E-Cell-Plessey Electronics) charged with a known amount of electricity. If the device is removed from the refrigerator, it begins to discharge at a known rate. The amount of time subsequently can be determined by the loss of charge. The prototype of this instrument has been found to be quite accurate and small (2 inches X 2 inches X 1 inch). It would be rather inexpensive if made in considerable numbers. PMID:867474

  19. Gas bubble disease monitoring and research of juvenile salmonids

    International Nuclear Information System (INIS)

    This document describes the project activities 1996--1997 contract year. This report is composed of three chapters which contain data and analyses of the three main elements of the project: field research to determine the vertical distribution of migrating juvenile salmonids, monitoring of juvenile migrants at dams on the Snake and Columbia rivers, and laboratory experiments to describe the progression of gas bubble disease signs leading to mortality. The major findings described in this report are: A miniature pressure-sensitive radio transmitter was found to be accurate and precise and, after compensation for water temperature, can be used to determine the depth of tagged-fish to within 0.32 m of the true depth (Chapter 1). Preliminary data from very few fish suggest that depth protects migrating juvenile steelhead from total dissolved gas supersaturation (Chapter 1). As in 1995, few fish had any signs of gas bubble disease, but it appeared that prevalence and severity increased as fish migrated downstream and in response to changing gas supersaturation (Chapter 2). It appeared to gas bubble disease was not a threat to migrating juvenile salmonids when total dissolved gas supersaturation was < 120% (Chapter 2). Laboratory studies suggest that external examinations are appropriate for determining the severity of gas bubble disease in juvenile salmonids (Chapter 3). The authors developed a new method for examining gill arches for intravascular bubbles by clamping the ventral aorta to reduce bleeding when arches were removed (Chapter 3). Despite an outbreak of bacterial kidney disease in the experimental fish, the data indicate that gas bubble disease is a progressive trauma that can be monitored (Chapter 3)

  20. Noninvasive 24-hour ambulatory arterial blood pressure monitoring in cirrhosis

    DEFF Research Database (Denmark)

    Møller, Søren; Wiinberg, N; Henriksen, Jens Henrik

    1995-01-01

    Cirrhotic patients have disturbed systemic hemodynamics with reduced arterial blood pressure, but this has not been investigated during daily activity and sleep. Systolic (SBP), diastolic (DBP), and mean arterial blood pressure (MAP), and heart rate (HR) were measured by an automatic ambulant...... were almost similar in the two groups (108 vs. 110; 65 vs. 67; 78 vs. 82 mm Hg, NS). Conversely, HR was significantly higher in the patients both in the daytime (86 vs. 72/min, P reduction in blood pressure and HR from daytime...

  1. Exposure-based treatment to control excessive blood glucose monitoring.

    OpenAIRE

    Allen, K D; Evans, J. H.

    2001-01-01

    We investigated an exposure-based procedure for reducing excessive checking of blood glucose by a child with diabetes. In a changing criterion design, an exposure-based procedure was implemented by systematically exposing the child to decreasing amounts of information about blood sugar levels (checking) and thereby increasing exposure to potential hypoglycemia. Access to information was reduced in graduated increments, with the parents setting criteria to levels at which they were willing to ...

  2. Reactive intermediates in the gas phase generation and monitoring

    CERN Document Server

    Setser, D W

    2013-01-01

    Reactive Intermediates in the Gas Phase: Generation and Monitoring covers methods for reactive intermediates in the gas phase. The book discusses the generation and measurement of atom and radical concentrations in flow systems; the high temperature flow tubes, generation and measurement of refractory species; and the electronically excited long-lived states of atoms and diatomic molecules in flow systems. The text also describes the production and detection of reactive species with lasers in static systems; the production of small positive ions in a mass spectrometer; and the discharge-excite

  3. Microfabricated BTU monitoring device for system-wide natural gas monitoring.

    Energy Technology Data Exchange (ETDEWEB)

    Einfeld, Wayne; Manginell, Ronald Paul; Robinson, Alex Lockwood; Moorman, Matthew Wallace

    2005-11-01

    The natural gas industry seeks inexpensive sensors and instrumentation to rapidly measure gas heating value in widely distributed locations. For gas pipelines, this will improve gas quality during transfer and blending, and will expedite accurate financial accounting. Industrial endusers will benefit through continuous feedback of physical gas properties to improve combustion efficiency during use. To meet this need, Sandia has developed a natural gas heating value monitoring instrument using existing and modified microfabricated components. The instrument consists of a silicon micro-fabricated gas chromatography column in conjunction with a catalytic micro-calorimeter sensor. A reference thermal conductivity sensor provides diagnostics and surety. This combination allows for continuous calorimetric determination with a 1 minute analysis time and 1.5 minute cycle time using air as a carrier gas. This system will find application at remote natural gas mining stations, pipeline switching and metering stations, turbine generators, and other industrial user sites. Microfabrication techniques will allow the analytical components to be manufactured in production quantities at a low per-unit cost.

  4. Continuous blood pressure monitoring in cirrhosis. Relations to splanchnic and systemic haemodynamics

    DEFF Research Database (Denmark)

    Møller, Søren; Christensen, E; Henriksen, Jens Henrik

    1997-01-01

    blood pressure and related to the results of an invasive haemodynamic investigation, including measurements of intra-arterial blood pressure (9.00-11.00 h) in 37 patients with cirrhosis. RESULTS: The 24-h blood pressures were significantly lower and the heart rate was significantly higher in patients......BACKGROUND/AIMS: Low arterial blood pressure is recognised as a distinctive factor in the hyperdynamic circulation in cirrhosis. 24-hour monitoring of the blood pressure and heart rate has recently revealed a reduced circadian variation with relation to liver function. However, associations...... with other clinical and haemodynamic characteristics have not been investigated and the aim of the present study was to identify splanchnic and systemic determinants of the 24-h blood pressure and heart rate in cirrhosis. METHODS: The variables were measured by an automatic ambulant device for monitoring...

  5. Evaluation of two methods of rapid blood-glucose monitoring by unskilled personnel during surgery

    DEFF Research Database (Denmark)

    Madsbad, S; Adelhøj, B; Bigler, Dennis Richard;

    1984-01-01

    The accuracy of two rapid methods of blood-glucose monitoring without (Haemo-glucotest 1-44) and with a reflectance meter (Hypocount B) was compared using a laboratory method. The assessment was carried out by personnel with no previous experience in measuring blood glucose. Eighty-five percent o...

  6. Self-Blood Glucose Monitoring in Diabetes Mellitus (Beyond the Basics)

    Science.gov (United States)

    ... Use ©2016 UpToDate, Inc. Patient education: Self-blood glucose monitoring in diabetes mellitus (Beyond the Basics) Author ... 2016. | This topic last updated: Oct 19, 2015. BLOOD SUGAR TESTING OVERVIEW — If you have diabetes, you have ...

  7. Novel Approach for Ensuring Increased Validity in Home Blood Pressure Monitoring

    DEFF Research Database (Denmark)

    Wagner, Stefan Rahr; Toftegaard, Thomas Skjødeberg; Bertelsen, Olav Wedege

    This paper proposes a novel technique to increase the validity of home blood pressure monitoring by using various sensor technologies as part of an intelligent environment platform in the home of the user. A range of recommendations exists on how to obtain a valid blood pressure...

  8. A SIGNAL ENHANCED PORTABLE RAMAN PROBE FOR ANESTHETIC GAS MONITORING

    Directory of Open Access Journals (Sweden)

    S. Schlüter

    2015-03-01

    Full Text Available The spontaneous Raman scattering technique is an excellent tool for a quantitative analysis of multi-species gas mixtures. It is a noninvasive optical method for species identification and gas phase concentration measurement of all Raman active molecules, since the intensity of the species specific Raman signal is linearly dependent on the concentration. Applying a continuous wave (CW laser it typically takes a few seconds to capture a gas phase Raman spectrum at room temperature. Nevertheless in contrast to these advantages the weak Raman signal intensity is a major drawback. Thus, it is still challenging to detect gas phase Raman spectra in alow-pressure regime with a temporal resolution of only a few 100 ms. In this work a fully functional gas phase Raman system for measurements in the low-pressure regime (p ≥ 980 hPa (absolute is presented. It overcomes the drawback of a weak Raman signal by using a multipass cavity. A description of the sensor setup and of the multipass arrangement will be presented. Moreover the complete functionality of the sensor system will be demonstrated by measurements at an anesthesia simulator under clinical relevant conditions and in comparison to a conventional gas monitor.

  9. Assessment of Self-Monitored Blood Glucose Results Using a Reflectance Meter with Memory and Microcomputer

    OpenAIRE

    Kuykendall, V.G.; Michaels, D W; Hartmann, K.G.

    1985-01-01

    A microcomputer software package for diabetes patient care utilizing self-monitoring of blood glucose (SMBG) has been developed. The software facilitates the collection, storage, analysis, and presentation of blood glucose/time information. Data entry is accomplished automatically via interface to hand held blood glucose reflectance instruments which retain up to 339 glucose/time results in internal RAM. The times of other significant clinical events may also be stored in the meter and upload...

  10. Noble gas geochemistry to monitor CO2 geological storages

    International Nuclear Information System (INIS)

    According to the last IPCC (Intergovernmental Panel on Climate Change) report, a probability of 90 % can be now established for the responsibility of the anthropogenic CO2 emissions for the global climate change observed since the beginning of the 20. century. To reduce these emissions and keep producing energy from coal, oil or gas combustions, CO2 could be stored in geological reservoirs like aquifers, coal beds, and depleted oil or gas fields. Storing CO2 in geological formations implies to control the efficiency and to survey the integrity of the storages, in order to be able to detect the possible leaks as fast as possible. Here, we study the feasibility of a geochemical monitoring through noble gas geochemistry. We present (1) the development of a new analytical line, Garodiox, developed to extract quantitatively noble gas from water samples, (2) the testing of Garodiox on samples from a natural CO2 storage analogue (Pavin lake, France) and (3) the results of a first field work on a natural CO2 accumulation (Montmiral, France). The results we obtain and the conclusions we draw, highlight the interest of the geochemical monitoring we suggest. (author)

  11. The physiological basis of pulmonary gas exchange: implications for clinical interpretation of arterial blood gases.

    Science.gov (United States)

    Wagner, Peter D

    2015-01-01

    The field of pulmonary gas exchange is mature, with the basic principles developed more than 60 years ago. Arterial blood gas measurements (tensions and concentrations of O₂ and CO₂) constitute a mainstay of clinical care to assess the degree of pulmonary gas exchange abnormality. However, the factors that dictate arterial blood gas values are often multifactorial and complex, with six different causes of hypoxaemia (inspiratory hypoxia, hypoventilation, ventilation/perfusion inequality, diffusion limitation, shunting and reduced mixed venous oxygenation) contributing variably to the arterial O₂ and CO₂ tension in any given patient. Blood gas values are then usually further affected by the body's abilities to compensate for gas exchange disturbances by three tactics (greater O₂ extraction, increasing ventilation and increasing cardiac output). This article explains the basic principles of gas exchange in health, mechanisms of altered gas exchange in disease, how the body compensates for abnormal gas exchange, and based on these principles, the tools available to interpret blood gas data and, quantitatively, to best understand the physiological state of each patient. This understanding is important because therapeutic intervention to improve abnormal gas exchange in any given patient needs to be based on the particular physiological mechanisms affecting gas exchange in that patient. PMID:25323225

  12. Skin perfusion pressure measured by isotope washout in legs with arterial occlusive disease. Evaluation of different tracers, comparison to segmental systolic pressure, angiography and transcutaneous oxygen tension and variations during changes in systemic blood pressure

    DEFF Research Database (Denmark)

    Holstein, P; Trap-Jensen, J; Bagger, H;

    1983-01-01

    digit (strain gauge technique). The two different methods correlated statistically significant at all four levels, but the systolic blood pressures were higher than the SPP in particular in diabetic legs; (3) Angiograms in 35 legs with AOD showed that the SPP on the ankle was only consistently decreased......The skin perfusion pressure (SPP) measured as the isotope washout cessation external pressure is valuable in selection of major amputation level. Five methodological investigations important to clinical use were carried out: (1) In five normal legs and 10 legs with arterial occlusive disease (AOD......Hg (range 18-98) (P less than 0.02). The average washout constant for the three different tracers were approximately equal and correlated statistically significant with the SPP; (2) In 59 legs with AOD, segmental SPP was compared to segmental systolic blood pressures on the thigh, calf, ankle and first...

  13. Extensive monitoring through multiple blood samples in professional soccer players

    DEFF Research Database (Denmark)

    Heisterberg, Mette F; Fahrenkrug, Jan; Krustrup, Peter;

    2013-01-01

    months period and analyzed for 37 variables in 27 professional soccer players from the best Danish league. Additionally, players were tested for body composition, VO2max and physical performance by the Yo-Yo intermittent endurance sub-max test (IE2).Multiple variations in blood parameters occurred during...

  14. Monitoring infection: from blood culture to polymerase chain reaction (PCR).

    Science.gov (United States)

    Book, Malte; Lehmann, Lutz Eric; Zhang, XiangHong; Stüber, Frank

    2013-06-01

    In patients with sepsis, diagnosis of blood stream infection (BSI) is a key concern to the therapist. Direct verification of pathogens in the blood stream executed by blood cultures (BC) still is regarded as the gold standard up to date. The quickest possible initiation of an appropriate antimicrobial therapy is a cornerstone of an effective therapy. Moreover, in this view BC can also serve to identify antimicrobial agents to target the pathogen. However, when employing BC the time needed until microbiological results are available ranges from 24 up to 72 h. Moreover, infections caused by multiple pathogens often remain undetected and concurrent antibiotic therapy may lower the overall sensitivity. Alternative pathogen characterization can be performed by polymerase chain reaction (PCR) based amplification methods. Results using PCR can be obtained within 6-8 h. Therefore, the time delay until an appropriate therapy can be reduced enormously. Moreover, these methods have the potential to enhance the sensitivity in the diagnosis of blood stream infections. Therefore, PCR based methods might be a valuable adjunct to present procedures of diagnosing bacteraemia.

  15. Long-term monitoring of marine gas leakage

    Science.gov (United States)

    Spickenbom, Kai; Faber, Eckhard; Poggenburg, Jürgen; Seeger, Christian; Furche, Markus

    2010-05-01

    The sequestration of CO2 in sub-seabed geological formations is one of the Carbon Capture and Storage (CCS) strategies currently under study. Although offshore operations are significantly more expensive than comparable onshore operations, the growing public resistance against onshore CCS projects makes sub-seabed storage a promising option. Even after a thorough review of the geological setting, there is always the possibility of leakage from the reservoir. As part of the EU-financed project CO2ReMoVe (Research, Monitoring, Verification), which aims to develop innovative research and technologies for monitoring and verification of carbon dioxide geological storage, we are working on the development of submarine long-term gas flow monitoring systems. The basic design of the monitoring system builds on our experience in volcano monitoring. Early prototypes were composed of a raft floating on the surface of a mud volcano, carrying sensors for CO2 flux and concentration, data storage and transmission, and power supply by battery-buffered solar panels. The system was modified for installation in open sea by using a buoy instead of a raft and a funnel on the seafloor to collect the gas, connected by a flexible tube. This setup provides a cost-effective solution for shallow waters. However, a buoy interferes with ship traffic, and it is also difficult to adapt this design to greater water depths. These requirements can best be complied by a completely submersed system. A system for unattended long-term monitoring in a marine environment has to be extremely durable. Therefore, we focussed on developing a mechanically and electrically as simple setup as possible, which has the additional advantage of low cost. The system consists of a funnel-shaped gas collector, a sensor head and pressure housings for electronics and power supply. Since this setup is inexpensive, it can be deployed in numbers to cover larger areas. By addition of multi-channel data loggers, data

  16. Long-term flow monitoring of submarine gas emanations

    Science.gov (United States)

    Spickenbom, K.; Faber, E.; Poggenburg, J.; Seeger, C.

    2009-04-01

    One of the Carbon Capture and Storage (CCS) strategies currently under study is the sequestration of CO2 in sub-seabed geological formations. Even after a thorough review of the geological setting, there is the possibility of leaks from the reservoirs. As part of the EU-financed project CO2ReMoVe (Research, Monitoring, Verification), which aims to develop innovative research and technologies for monitoring and verification of carbon dioxide geological storage, we are working on the development of submarine long-term gas flow monitoring systems. Technically, however, these systems are not limited to CO2 but can be used for monitoring of any free gas emission (bubbles) on the seafloor. The basic design of the gas flow sensor system was derived from former prototypes developed for monitoring CO2 and CH4 on mud volcanoes in Azerbaijan. This design was composed of a raft floating on the surface above the gas vent to collect the bubbles. Sensors for CO2 flux and concentration and electronics for data storage and transmission were mounted on the raft, together with battery-buffered solar panels for power supply. The system was modified for installation in open sea by using a buoy instead of a raft and a funnel on the seafloor to collect the gas, which is then guided above water level through a flexible tube. Besides some technical problems (condensed water in the tube, movement of the buoys due to waves leading to biased measurement of flow rates), this setup provides a cost-effective solution for shallow waters. However, a buoy interferes with ship traffic, and it is also difficult to adapt this design to greater water depths. These requirements can best be complied by a completely submersed system. To allow unattended long-term monitoring in a submarine environment, such a system has to be extremely durable. Therefore, we focussed on developing a mechanically and electrically as simple setup as possible, which has the additional advantage of low cost. The system

  17. Optical coherence tomography for glucose monitoring in blood

    Science.gov (United States)

    Ullah, Hafeez; Hussain, Fayyaz; Ikram, Masroor

    2015-08-01

    In this review, we have discussed the potential application of the emerging imaging modality, i.e., optical coherence tomography (OCT) for glucose monitoring in biological tissues. OCT provides monitoring of glucose diffusion in different fibrous tissues like in sclera by determining the permeability rate with acceptable accuracy both in type 1 and in type 2 diabetes. The maximum precision of glucose measurement in Intralipid suspensions, for example, with the OCT technique yields the accuracy up to 4.4 mM for 10 % Intralipid and 2.2 mM for 3 % Intralipid.

  18. Continuous monitoring of blood glucose in the practice of endocrinologist

    Directory of Open Access Journals (Sweden)

    Yu I Philippov

    2012-12-01

    Full Text Available Continuous glucose monitoring - an important diagnostic, teaching and treatment tool for patients with diabetes mellitus, which is increasingly becoming a part of routine clinical practice in endocrinology. This article presents an overview of modern techniques, their advantages and disadvantages, evidence basis and place in everyday clinical practice. The article discusses the key factors affecting the efficiency, indications, contraindications, conditions of use of the continuous glucose monitoring systems in patients with diabetes mellitus, gives an algorithm for the application of this technique in clinical practice.

  19. Cytokine signatures of human whole blood for monitoring immunosuppression

    OpenAIRE

    He, Yi; Luo, Yuwei; Lao, Xiaobin; Tan, Liping; Sun, Erwei

    2014-01-01

    How to evaluate status of the immune system is extremely critical for clinical immunosuppressive treatment. In this study, we tested the secretion of cytokines in undiluted whole blood samples stimulated with Phorbol 12-myristate 13-acetate (PMA) and ionomycin (IONO), and compared the effects of dexamethasone (DEX), cyclosporine A (CsA) or mycophenolic acid (MPA), either alone or in combination, on cytokine profiles. The results showed that both DEX and CsA dose-dependently inhibited the prod...

  20. On preventive blood pressure self-monitoring at home

    DEFF Research Database (Denmark)

    Verdezoto, Nervo; Grönvall, Erik

    2015-01-01

    to develop better tools to support people’s preventive self-monitoring needs, compared to existing solutions. Supporting the active and informed individual can help improve people’s self-care, awareness, and implementation of preventive care. Based on our study, we also reflect on the findings to illustrate...

  1. Monitoring oxygenation.

    Science.gov (United States)

    Severinghaus, John W

    2011-06-01

    Cyanosis was used for a century after dentists began pulling teeth under 100% N(2)O in 1844 because brief (2 min) severe hypoxia is harmless. Deaths came with curare and potent anesthetic respiratory arrest. Leland Clark's invention of a polarographic blood oxygen tension electrode (1954) was introduced for transcutaneous PO2 monitoring to adjust PEEP and CPAP PO2 to prevent premature infant blindness from excess O2 (1972). Oximetry for warning military aviators was tried after WW II but not used for routine monitoring until Takuo Aoyagi (1973) discovered an equation to measure SaO2 by the ratio of ratios of red and IR light transmitted through tissue as it changed with arterial pulses. Pulse oximetry (1982) depended on simultaneous technology improvements of light emitting red and IR diodes, tiny cheap solid state sensors and micro-chip computers. Continuous monitoring of airway anesthetic concentration and oxygen also became very common after 1980. Death from anesthesia fell 10 fold between 1985 and 2000 as pulse oximetry became universally used, but no proof of a causative relationship to pulse oximetry exists. It is now assumed that all anesthesiologist became much more aware of the dangers of prolonged hypoxia, perhaps by using the pulse oximeters. PMID:21717228

  2. The latest continuous monitoring instrumentation for ground-gas monitoring and risk prediction

    Directory of Open Access Journals (Sweden)

    A. N. Nwachukwu, A. W. Diya

    2012-01-01

    Full Text Available In the field of ground-gas monitoring and risk assessment, the concentrations and flows of ground-gas from contaminated sites are usually measured in-situ using handheld or transportable field instruments followed with ex-situ analysis. These portable instruments only give spot data of the gases being monitored and few of their controls. However, and more recently, it has been discovered that ground-gas concentrations and flows are temporarily and spatially variable, with the accompanying realization that the existing instruments are not able to resolve these problems. For example, contaminated sites investigation shows that environment conditions such as pressure, temperature, water level, and air movement substantially affect ground-gas concentrations on the range of timescales, therefore; uncertainty will always exist when using methods that lack temporal resolution. This paper describes the most recent instrumentation that provides real time continuous gas data both inside borehole and ambient air for parameters such as methane, carbon dioxide, volatile organic compounds (VOCs, hydrogen sulfides, oxygen and carbon monoxide alongside atmospheric and borehole pressures. Raw data is presented from field work carried out in 2 landfill sites and 2 peat sites in Manchester, United Kingdom. These datasets are currently being used to derive a methodology for improved prediction of risks due to ground-gases.

  3. Transcutaneous electrostimulation for osteoarthritis of the knee: CAT

    Directory of Open Access Journals (Sweden)

    Raúl Alberto Aguilera Eguía

    2013-09-01

    Full Text Available Purpose. The aim of this CAT (Critically Appraised Topic was to check the validity of the results and effectiveness of Transcutaneous Electrical Stimulation in subjects with knee osteoarthritis and answer the question: In subjects with osteoarthritis of the knee, does low frequency transcutaneous electrical stimulation reduce pain? Method. We conducted an analysis of the article "Transcutaneous Electrical Stimulation for osteoarthritis of the knee, Cochrane Systematic Review" of Rutjes et al (2009. We analyzed validity of results, applicability and effectiveness of this intervention in patients with knee osteoarthritis. Results. Low frequency transcutaneous electrical stimulation may reduce pain in subjects with knee osteoarthritis. SMD -0.85 (95% CI -1.36 to -0.34. Conclusion. The intervention is neither endorsed nor discouraged in knee pain reduction.

  4. Transcutaneous immunisation assisted by low-frequency ultrasound.

    OpenAIRE

    Dahlan, A.; Alpar, H. O.; Stickings, P.; Sesardic, D.; Murdan, S.

    2009-01-01

    Low-frequency ultrasound application is known to increase the skin's permeability to large molecules such as vaccines, and to enable transcutaneous immunisation. Sodium dodecyl sulphate (SDS) - a skin irritant - is often included in the coupling medium at 1% (w/v), as this has been found to enhance skin permeability. In this paper we show, for the first time, the feasibility of low-frequency ultrasound-assisted transcutaneous immunisation in the absence of SDS. Antibody titres were strongly i...

  5. Transcutaneous electrical nerve stimulation (TENS) in angina pectoris.

    Science.gov (United States)

    Mannheimer, C; Carlsson, C A; Vedin, A; Wilhelmsson, C

    1986-09-01

    The aim of this study was to determine the efficacy of transcutaneous electrical nerve stimulation (TENS) in the treatment of chronic stable severe angina pectoris. In a short-term study the effect of TENS was studied in 10 male patients with angina pectoris (functional class III and IV). All patients had previously been stabilized on long-term maximal oral treatment. The effects of the treatment were measured by means of repeated bicycle ergometer tests. All patients had an increased working capacity (16-85%), decreased ST segment depression and reduced recovery time during TENS. No adverse effects were observed. A long-term study of TENS on similarly selected patients showed beneficial effects in terms of pain reduction, reduced frequency of anginal attacks, increased physical activity and increased working capacity during bicycle ergometer tests. An invasive study was carried out with respect to systemic and coronary hemodynamics and myocardial metabolism during pacing provoked myocardial ischemia in 13 patients. The results showed that TENS led to an increased tolerance to pacing, improved lactate metabolism, less pronounced ST segment depression. A drop in systolic blood pressure during TENS treatment at identical pacing rates indicated a decreased afterload. An increased coronary flow to ischemic areas in the myocardium was supported by the fact that the rate pressure product during anginal pain increased during TENS.

  6. Noninvasive monitoring of peripheral microcirculatory hemodynamics under varying degrees of hypoxia.

    Science.gov (United States)

    Ovadia-Blechman, Zehava; Meilin, Aviram; Rabin, Neta; Eldar, Michael; Castel, David

    2015-09-15

    The effect of hypoxia on skin blood flow was examined in anesthetized rabbits during induction of various levels of hypoxia. Peripheral perfusion and oxygenation were monitoring using a combined system (LPT) composed of a laser Doppler flowmeter (LDF), a photoplatysmograph (PPG), and a transcutaneous oxygen tension monitor (tc-PO2). Central blood parameters (PaO2, HCO3(-), SaO2, pH, and lactate) were measured concomitantly throughout the experiment. A continuous decline was found in both peripheral and central values, depending on the severity of the hypoxia. The results clearly indicate that monitoring peripheral indices with the LPT system enables monitoring changes of vital blood parameters during hypoxia. The system has clinical potential for sensitive and noninvasive monitoring of vital variables during medical procedures in clinics, as well as for homecare for patients with respiratory diseases. Minimizing the system may be useful in various conditions of exposure to low oxygen levels, such as during mountain climbing.

  7. Transcutaneous optical power converter for implantable devices

    Science.gov (United States)

    Tamura, Toshiyo; Shamsuddin, A. K. M.; Kawarada, Atsushi; Togawa, Tatsuo; Oberg, P. Ake

    1994-02-01

    An optical transcutaneous power converter has been developed for the power supply of implanted devices. It consists of a light source, optical fiber system and a photo detector. The light source is either a halogen lamp or continuous high-power laser diode which illuminated skin surface. The light penetrates through the skin to the solar cells. The preliminary experiments with a slice of tissue in between the light source and the solar cells showed that the power transfer efficiency was 40% in comparison to direct illumination of the solar cells. The maximum electric power obtained with a slice of tissue simulating the skin was about 8 mW at a laser diode run at a power of 100 mW. The electric power transferred is enough to supply low power consuming CMOS circuits.

  8. Monitoring

    Science.gov (United States)

    ... its main source of fuel. To keep your blood sugar level on target and avoid problems with your eyes, kidneys, heart and feet, you should eat right ... better. And monitoring doesn’t stop at measuring blood sugar levels. Because ... blood testing) Eye health (eye exams) Foot health (foot exams and ...

  9. Lamotrigine in pregnancy - therapeutic drug monitoring in maternal blood, amniotic fluid, and cord blood.

    Science.gov (United States)

    Paulzen, Michael; Lammertz, Sarah E; Veselinovic, Tanja; Goecke, Tamme W; Hiemke, Christoph; Gründer, Gerhard

    2015-09-01

    This study is the first to measure and correlate lamotrigine concentrations in maternal blood, amniotic fluid, and umbilical cord blood and account for distribution of the drug between these three compartments. Concentrations of lamotrigine were measured in six mother-infant pairs at the time of delivery. Daily doses of lamotrigine ranged between 200 and 650 mg. Daily doses were correlated with maternal serum and umbilical cord blood concentrations, and serum levels were correlated with levels in amniotic fluid. Lamotrigine levels in serum correlated strongly with the lamotrigine levels in amniotic fluid (r=+0.986, Pamniotic fluid was in a range between 0.31 and 0.75 (mean 0.58, SD 0.17); the penetration ratio into the fetal circulation, calculated on the basis of umbilical cord blood levels, was found to be in a range between 0.48 and 1.27 (mean 0.81, SD 0.28). Lamotrigine concentrations in amniotic fluid provided evidence that maternally administered lamotrigine is accessible to the fetus in a manner not previously appreciated. Furthermore, the penetration ratio into umbilical cord blood calculated here is in line with the largest study carried out so far to explore transplacental transfer.

  10. Development of monitoring and control technology based on trace gas monitoring. Final report

    Energy Technology Data Exchange (ETDEWEB)

    Liebowitz, B.

    1997-07-01

    Trace gases are generated by many biological reactions. During anaerobic decomposition, trace levels of hydrogen (H{sub 2}) and carbon monoxide (CO) gases are produced. It was shown previously that these trace gases are intrinsically related to the biochemical reactions occurring and, therefore, offer promise for on-line process monitoring and control. This work was designed to test how effectively hydrogen and CO could be to monitor high-rate anaerobic systems that has significant mass transfer and complex hydraulics. An experimental program was designed to examine the behavior of an upflow anaerobic sludge blanket (UASB) reactor system under steady state and in response to organic loading perturbations. The responses of trace gases CO and H{sub 2} were tracked using an on-line, real-time gas-monitoring system linked to a computer-controlled data acquisition package. Data on conventional process parameters such as pH, chemical oxygen demand (COD), volatile fatty acids (VFAs) were concurrently collected. Monitoring of conventional process indicators (i.e., pH, VFA, gas production) and trace gas (H{sub 2} and CO) indicators was conducted using a matrix of nine different steady-state OLRs (4-23 kg COD/m{sup 3} -d) and system HRTs (0.5 to 2.5 days) was performed to determine any correlation among the indicators. Of OLR, HRT, and influent COD, only OLR had any significant influence on the process indicators examined. All parameters except methane increased with increases in OLR; methane decreased with increased OLR. The OLR and gas production rate (GP) were observed to be linearly correlated.

  11. Beam Impedance Studies of the PS Beam Gas Ionization Monitor

    CERN Document Server

    Avgidis, Fotios

    2016-01-01

    The Beam Gas Ionization monitor (BGI) is a device for continuous beam size monitoring that is intended to be installed in the CERN Proton Synchrotron (PS) during the extended year-end technical stop from December 2016 to April 2017. With the objective of determining the impedance contribution of the BGI vacuum chamber to the overall beam impedance, we report on RF measurements on the device in a laboratory frame, measurement data analysis, and RF simulations of the structure under investigation. For the impedance contribution characterization of the BGI, the following approach has been followed: First, the EM fields inside a simplified BGI model that doesn’t include any of the internal components of the vacuum chamber have been simulated. RF measurements have been performed on the same empty structure showing great agreement between measurement and simulation and thus verifying the validity of the model. Second, simulations have been executed on a fully assembled BGI model that includes all the internal ele...

  12. Laser Spectroscopy Based Multi-Gas Monitor Technology Demonstration

    Science.gov (United States)

    Mudgett, Paul D.; Pilgrim, Jeffrey S.

    2016-01-01

    The timing was right in the “evolution” of low power tunable diode laser spectroscopy (TDLS) to design a spacecraft cabin air monitor around technology being developed at a small company funded by SBIR grants. NASA Centers had been monitoring their progress hoping that certain key gaps in the long term gas monitoring development roadmap could be filled by TDLS. The first iteration of a monitor for multiple gases called the Multi-Gas Monitor (MGM) which measures oxygen, carbon dioxide, ammonia and water vapor, as well as temperature and pressure. In January 2013, the ISS Program being particularly interested in ammonia funded a technology demonstration of MGM. The project was a joint effort between Vista Photonics for the sensor, NASA-JSC for project management and laboratory calibration, and Nanoracks for the enclosure and payload certification/integration. Nanoracks was selected in order to use their new experimental infrastructure located in an EXPRESS rack in the JEM. The MGM enclosure has multiple power supply options including 5VDC USB interface to the Nanoracks Frame, 28VDC Express Rack power and internal rechargeable batteries. MGM was calibrated at NASA-JSC in July 2013, delivered to ISS on 37 Soyuz in November 2013 and was installed and activated in February 2014. MGM resided in the Nanoracks Frame making continuous measurements the majority of the time, but also spent a day in Node 3 on battery power, and a month in the US Lab Module on 28VDC power, as part of the demonstration. Data was downloaded via Nanoracks on roughly a weekly basis. Comparisons were made with data from the Major Constituents Analyzer (MCA) which draws and analyzes air from JEM and other modules several times per hour. A crewmember challenged the carbon dioxide channel by breathing into the intake upon startup, and challenged the ammonia channel later using a commercial ammonia inhalant. Many interesting phenomena in the cabin atmosphere were detected during the tech demo

  13. Continuous monitoring of a changing sample by multiplex gas chromatography

    Science.gov (United States)

    Valentin, Jose R.; Hall, Kirsten W.; Becker, Joseph F.

    1990-01-01

    Results are presented from a study in which a continuously changed gaseous sample was monitored by multiplex gas chromatography (MGC), using the exponential dilution (ED) technique of Ritter and Adams (1976) to change the composition and concentration of a gaseous mixture in such a way as to imitate changes in the atmospheric gases sampled by a descending aircraft. A calibration of the MGC system was performed with four different rates of sample dilution, and the errors resulting from various degrees of change in the sample concentration were determined.

  14. Noninvasive cerebral blood oxygenation monitoring: clinical test of multiwavelength optoacoustic system

    Science.gov (United States)

    Petrov, Y. Y.; Prough, D. S.; Petrova, I.; Patrikeev, I. A.; Cicenaite, I.; Esenaliev, R. O.

    2007-02-01

    Continuous monitoring of cerebral blood oxygenation is critically important for treatment of patients with life-threatening conditions like severe brain injury or during cardiac surgery. We designed and built a novel multiwavelength optoacoustic system for noninvasive, continuous, and accurate monitoring of cerebral blood oxygenation. We use an Optical Parametric Oscillator as a light source. We successfully tested the system in vitro as well as in vivo in large animals (sheep) through thick tissues overlying blood vessels which drain venous blood out of the brain (e.g., superior sagittal sinus or jugular vein). Here we present the results of clinical tests of the system for continuous noninvasive cerebral blood oxygenation monitoring in the internal jugular vein of healthy volunteers. We applied our custom-built optoacoustic probe (which incorporated a wide-band acoustic transducer and an optical fiber) to the neck area overlying the internal jugular vein. We performed measurements with volunteers at 18 wavelengths in the near-infrared spectral range. Despite a thick layer of overlying connective tissue and low energy used in the experiments, we recorded signals with high signal-to-noise ratios for all volunteers. We found that the temporal (independent of signal amplitude) parameters of recorded profiles for different levels of blood oxygenation correlated well with the spectrum of effective attenuation coefficients of blood.

  15. Arterial blood gas analysis or oxygen saturation in the assessment of acute asthma?

    OpenAIRE

    Carruthers, D M; Harrison, B. D.

    1995-01-01

    BACKGROUND--A study was undertaken to determine if arterial blood gas estimation is always necessary in the assessment of patients presenting to hospital with acute severe asthma, or whether oxygen saturation as measured by pulse oximetry is a reliable screening test for predicting those in respiratory failure. METHODS--A prospective study was conducted in a specialist respiratory medical unit. Arterial blood gas tensions and pulse oximetry were measured in 89 consecutive patients admitted wi...

  16. Non-invasive respiratory monitoring in paediatric intensive care unit.

    Directory of Open Access Journals (Sweden)

    Nadkarni U

    2000-04-01

    Full Text Available Monitoring respiratory function is important in a Paediatrics Intensive Care Unit (PICU, as majority of patients have cardio-respiratory problems. Non-invasive monitoring is convenient, accurate, and has minimal complications. Along with clinical monitoring, oxygen saturation using pulse oximetry, transcutaneous oxygenation (PtcO2 and transcutaneous PCO2 (PtcCO2 using transcutaneous monitors and end-tidal CO2 using capnography are important and routine measurements done in most PICUs. Considering the financial and maintenance constraints pulse oximetry with end tidal CO2 monitoring can be considered as most feasible.

  17. Continuous non-invasive finger blood pressure monitoring in children.

    Science.gov (United States)

    Tanaka, H; Thulesius, O; Yamaguchi, H; Mino, M; Konishi, K

    1994-06-01

    We evaluated the performance of continuous non-invasive finger arterial pressure measurement using the volume-clamp technique (Finapres). This study was designed to compare finger arterial pressure with brachial blood pressure estimated by the auscultatory method in 217 children (90 boys and 127 girls) aged 4-16 years and in 38 adults (aged 18-45 years). Finger and brachial artery pressure readings were obtained consecutively from the ipsilateral side in the supine position. Finger arterial pressure waveforms were recorded in all children except 4 with small and thin fingers. There was good agreement for systolic pressure with only a slight underestimation of 1.9 mmHg and 5.1 mmHg lower for diastolic pressure. This difference most probably reflects inaccuracy of the auscultatory cuff method rather than an error in the Finapres. There was large inter-individual variability in Finapres recordings which might be due to differences in vasomotor tone, as demonstrated by systolic amplification in 5 patients with anorexia. However, Finapres showed a small within-subject variability (3.8 mmHg for systolic and 4.1 mmHg for diastolic pressure) determined in 5 patients during phenylephrine infusion, and as good reproducibility as the auscultatory method. These results suggest that finger arterial pressure measurement in children older than 6 years of age has similar accuracy as that in adults, and that this method is useful for clinical applications in children, especially for the non-invasive evaluation of autonomic control and cardiovascular reflexes involving transient and rapid blood pressure changes. PMID:7919764

  18. Continuous blood pressure monitoring in cirrhosis. Relations to splanchnic and systemic haemodynamics

    DEFF Research Database (Denmark)

    Møller, S; Christensen, E; Henriksen, Jens Henrik Sahl

    1997-01-01

    a high post-sinusoidal resistance, a low plasma volume, a short central circulation time, and the presence of ascites. In contrast, a low intra-arterial blood pressure was determined by a low serum sodium, a low haemoglobin, and a high cardiac output. Diuretic treatment did not influence this model......BACKGROUND/AIMS: Low arterial blood pressure is recognised as a distinctive factor in the hyperdynamic circulation in cirrhosis. 24-hour monitoring of the blood pressure and heart rate has recently revealed a reduced circadian variation with relation to liver function. However, associations with...... other clinical and haemodynamic characteristics have not been investigated and the aim of the present study was to identify splanchnic and systemic determinants of the 24-h blood pressure and heart rate in cirrhosis. METHODS: The variables were measured by an automatic ambulant device for monitoring...

  19. Importance of monitoring family members in establishing sources and pathways of lead in blood.

    Science.gov (United States)

    Gulson, B L; Mizon, K J; Korsch, M J; Howarth, D

    1996-10-11

    High precision lead isotope measurements were undertaken to establish the sources and pathways in blood and environmental samples of five families from the Broken Hill lead mining community, New South Wales, Australia. The five families were selected from 27 families investigated to illustrate the different sources and pathways of lead into blood and the importance of monitoring the whole family. The results illustrate that although the major source of lead is from the orebody, paint and petrol can be significant contributors to both house dust and blood leads. The results also show that the sources and pathways can be from the father's occupation and hence monitoring of families is important, especially in high risk locations. In two cases, the elevated blood leads in the children did not derive from their current residence but from other residences in the community.

  20. The Health Behavior Schedule-II for Diabetes Predicts Self-Monitoring of Blood Glucose

    Science.gov (United States)

    Frank, Maxwell T.; Cho, Sungkun; Heiby, Elaine M.; Lee, Chun-I; Lahtela, Adrienne L.

    2006-01-01

    The Health Behavior Schedule-II for Diabetes (HBS-IID) is a 27-item questionnaire that was evaluated as a predictor of self-monitoring of blood glucose (SMBG). The HBS-IID was completed by 96 adults with Type 2 diabetes. Recent glycosylated hemoglobin HbA1c and fasting blood glucose results were taken from participants' medical records. Only 31.3%…

  1. Microvascular blood flow monitoring with laser speckle contrast imaging using the generalized differences algorithm

    OpenAIRE

    Humeau-Heurtier, Anne; Mahé, Guillaume; Abraham, Pierre

    2015-01-01

    Laser speckle contrast imaging (LSCI) is a full-field optical technique to monitor microvascular blood flow with high spatial and temporal resolutions. It is used in many medical fields such as dermatology, vascular medicine, or neurosciences. However, LSCI leads to a large amount of data: image sampling frequency is often of several Hz and recordings usually last several minutes. Therefore, clinicians often perform regions of interest in which a spatial averaging of blood flow is performed a...

  2. Clinical implication of blood glucose monitoring in general dental offices: the Ehime Dental Diabetes Study

    OpenAIRE

    Harase, Tadahiro; Nishida, Wataru; Hamakawa, Tomohiro; Hino, Satoshi; Shigematsu, Kenji; Kobayashi, Satoru; Sako, Hirofumi; Ito, Shirou; Murakami, Hajime; Nishida, Kei; Inoue, Hiroshi; Fujisawa, Masahito; Yoshizu, Hiroshi; Kawamura, Ryoichi; Takata, Yasunori

    2015-01-01

    Objective We examined whether general dentists can contribute to the detection of patients with undiagnosed diabetes and prediabetes by monitoring blood glucose in dental clinics. Research design and methods A total of 716 patients who visited clinics for dental treatment were enrolled and classified into 3 groups (mild, moderate, and severe) according to Kornman's criteria for periodontitis. The correlations between the casual blood glucose level, presence or absence of the history of diabet...

  3. System Accuracy Evaluation of the GlucoRx Nexus Voice TD-4280 Blood Glucose Monitoring System

    OpenAIRE

    Muhammad Khan; Keith Broadbent; Mike Morris; David Ewins; Franklin Joseph

    2014-01-01

    Use of blood glucose (BG) meters in the self-monitoring of blood glucose (SMBG) significantly lowers the risk of diabetic complications. With several BG meters now commercially available, the International Organization for Standardization (ISO) ensures that each BG meter conforms to a set degree of accuracy. Although adherence to ISO guidelines is a prerequisite for commercialization in Europe, several BG meters claim to meet the ISO guidelines yet fail to do so on internal validation. We con...

  4. Blood gas analyzer utility in evaluating oxygen kinetics of the aqueous humor

    Directory of Open Access Journals (Sweden)

    Ismail Ersan

    2015-04-01

    Full Text Available Purpose: To measure the partial pressure of oxygen (PO2 and carbon dioxide (PCO2 and the pH of aqueous humor (AH and arterial blood samples from rabbits using a blood gas analyzer. Methods: Twenty New Zealand rabbits were anesthetized intramuscularly with ketamine and xylazine and were then allowed to breathe room air. Using a gas blood analyzer, arterial blood and AH samples were analyzed for PO2, PCO2, and pH. Results: The mean arterial blood pressure was 87.14 ± 15.0 mmHg. The mean blood and AH PO2 were 95.18 ± 11.76 mmHg and 88.83 ± 9.92 mmHg, the mean blood and AH PCO2 were 25.86 ± 5.46 mmHg and 29.50 ± 5.36 mmHg, and the mean blood and AH pH were 7.38 ± 0.06 and 7.33 ± 0.09, respectively. Conclusion: Conclusions: The blood gas analyzer was easily employed to evaluate the aqueous humor in rabbits. When comparing the results of studies evaluating aqueous PO2, care should be taken to determine the methods used in these studies.

  5. Novel optoacoustic system for noninvasive continuous monitoring of cerebral venous blood oxygenation

    Science.gov (United States)

    Petrov, Yuriy; Petrov, Irene Y.; Prough, Donald S.; Esenaliev, Rinat O.

    2012-02-01

    Traumatic brain injury (TBI) and spinal cord injury are a major cause of death for individuals under 50 years of age. In the USA alone, 150,000 patients per year suffer moderate or severe TBI. Moreover, TBI is a major cause of combatrelated death. Monitoring of cerebral venous blood oxygenation is critically important for management of TBI patients because cerebral venous blood oxygenation below 50% results in death or severe neurologic complications. At present, there is no technique for noninvasive, accurate monitoring of this clinically important variable. We proposed to use optoacoustic technique for noninvasive monitoring of cerebral venous blood oxygenation by probing cerebral veins such as the superior sagittal sinus (SSS) and validated it in animal studies. In this work, we developed a novel, medical grade optoacoustic system for continuous, real-time cerebral venous blood oxygenation monitoring and tested it in human subjects at normal conditions and during hyperventilation to simulate changes that may occur in patients with TBI. We designed and built a highly-sensitive optoacoustic probe for SSS signal detection. Continuous measurements were performed in the near infrared spectral range and the SSS oxygenation absolute values were automatically calculated in real time using a special algorithm developed by our group. Continuous measurements performed at normal conditions and during hyperventilation demonstrated that hyperventilation resulted in approximately 12% decrease of cerebral venous blood oxygenation.

  6. Risk Stratification by Ambulatory Blood Pressure Monitoring Across JNC Classes of Conventional Blood Pressure

    DEFF Research Database (Denmark)

    Brguljan-Hitij, Jana; Thijs, Lutgarde; Li, Yan;

    2014-01-01

    BACKGROUND: Guidelines propose classification of conventional blood pressure (CBP) into normotension (hypertension (≥140/≥90 mm Hg). METHODS: To assess the potential differential contribution of ambulatory blood pressure (ABP...... with normotension (n = 2,639), prehypertension (n = 3,076) carried higher risk (P ≤ 0.015) of cardiovascular (+41%) and cerebrovascular (+92%) endpoints; compared with hypertension (n = 2,111) prehypertension entailed lower risk (P ≤ 0.005) of total mortality (-14%) and cardiovascular mortality (-29......%) and of cardiovascular (-34%), cardiac (-33%), or cerebrovascular (-47%) events. Multivariable-adjusted hazard ratios (HRs) for stroke associated with 24-hour and daytime diastolic ABP (+5 mm Hg) were higher (P ≤ 0.045) in normotension than in prehypertension and hypertension (1.98 vs.1.19 vs.1.28 and 1.73 vs.1.09 vs. 1...

  7. Comparison of invasive and non-invasive blood pressure monitoring during clinical anaesthesia in dogs.

    Science.gov (United States)

    MacFarlane, Paul D; Grint, Nicola; Dugdale, Alexandra

    2010-03-01

    Monitoring blood pressure during anaesthesia is widely recommended in man and animals. The accuracy of any device used to measure blood pressure is an important consideration when selecting monitoring equipment, the ANSI/AAMI SP10 standard is widely cited in this respect in recent veterinary publications. Blood pressure was monitored using invasive and non-invasive techniques during clinical anaesthesia in 19 dogs. The results were compared using Bland-Altman analysis. The bias (and limits of agreement) between invasive and non-invasive measurement was 7.1 mmHg (+/-34.7) for systolic blood pressure, -1.8 mmHg (+/-27.4) for mean blood pressure and 6.9 mmHg (+/-27.5) for diastolic blood pressure. In a clinical setting the bias between invasive and non-invasive measurement techniques was similar or smaller than laboratory reports, however the limits of agreement were considerably wider suggesting that care should be exercised when interpreting NIBP values. PMID:20306347

  8. Direct monitoring of organic vapours with amperometric enzyme gas sensors.

    Science.gov (United States)

    Hämmerle, Martin; Hilgert, Karin; Achmann, Sabine; Moos, Ralf

    2010-02-15

    In this study, amperometric enzyme gas sensors for direct monitoring of organic vapours (formaldehyde, ethanol and phenol) are presented using exemplarily different sensing strategies: NADH detection, H(2)O(2) detection and direct substrate recycling, respectively. The presented sensor configurations allow the selective, continuous, online monitoring of organic vapours without prior accumulation or sampling of the analyte. The gaseous samples are provided as headspace above aqueous solutions. The concentration in the gas phase was calculated from the concentration in solution at room temperature according to the respective Henry constants given in the literature. The enzymes employed are NAD-dependent formaldehyde dehydrogenase [EC 1.2.1.46] from Pseudomonas putida, alcohol oxidase [EC 1.1.3.13] from Pichia pastoris, and tyrosinase [EC 1.14.18.1] from mushroom. The gas diffusion working electrodes used in the sensors are based on a porous, hydrophobic PTFE membrane (exposed geometric electrode area: 1.77 cm(2)) covered with a porous layer of gold, platinum or graphite/Teflon. Detection limit, sensitivity, and measuring range are 34 microM (6.5 ppb), 117 nA/mM, and 0.46-66.4 mM for formaldehyde, 9.9 microM (55 ppb), 3.43 microA/mM, and 0.1-30 mM for ethanol, and 0.89 microM (0.36 ppb), 2.4 microA/mM, and 0.01-1 mM for phenol, respectively. Further sensor characteristics such as response time and stability are also determined: t(90%) (formaldehyde: 4.5 min; ethanol: 69 s; phenol: 27 min), stability at permanent exposure (formaldehyde: 63%, 15 h @ 2.62 mM; ethanol: 86%, 18 @ 1 mM; phenol: 86%, 16.5 h @ 0.1 M). PMID:19926472

  9. Condition Based Monitoring of Gas Turbine Combustion Components

    Energy Technology Data Exchange (ETDEWEB)

    Ulerich, Nancy; Kidane, Getnet; Spiegelberg, Christine; Tevs, Nikolai

    2012-09-30

    The objective of this program is to develop sensors that allow condition based monitoring of critical combustion parts of gas turbines. Siemens teamed with innovative, small companies that were developing sensor concepts that could monitor wearing and cracking of hot turbine parts. A magnetic crack monitoring sensor concept developed by JENTEK Sensors, Inc. was evaluated in laboratory tests. Designs for engine application were evaluated. The inability to develop a robust lead wire to transmit the signal long distances resulted in a discontinuation of this concept. An optical wear sensor concept proposed by K Sciences GP, LLC was tested in proof-of concept testing. The sensor concept depended, however, on optical fiber tips wearing with the loaded part. The fiber tip wear resulted in too much optical input variability; the sensor could not provide adequate stability for measurement. Siemens developed an alternative optical wear sensor approach that used a commercial PHILTEC, Inc. optical gap sensor with an optical spacer to remove fibers from the wearing surface. The gap sensor measured the length of the wearing spacer to follow loaded part wear. This optical wear sensor was developed to a Technology Readiness Level (TRL) of 5. It was validated in lab tests and installed on a floating transition seal in an F-Class gas turbine. Laboratory tests indicate that the concept can measure wear on loaded parts at temperatures up to 800{degrees}C with uncertainty of < 0.3 mm. Testing in an F-Class engine installation showed that the optical spacer wore with the wearing part. The electro-optics box located outside the engine enclosure survived the engine enclosure environment. The fiber optic cable and the optical spacer, however, both degraded after about 100 operating hours, impacting the signal analysis.

  10. Influence of different storage times and temperatures on blood gas and acid-base balance in ovine venous blood

    Directory of Open Access Journals (Sweden)

    H.A. Hussein

    2013-01-01

    Full Text Available The present study was designed to investigate the effects of storage temperature and time on blood gas and acid-base balance of ovine venous blood. Ten clinically healthy sheep were used in this study. A total number of 30 blood samples, were divided into three different groups, and were stored in a refrigerator adjusted to +4 ºC (Group I, n = 10, at RT of about 22-25 ºC (Group II, n = 10 and in an incubator adjusted to 37 ºC (Group III, n = 10 for up to 48 h. Blood samples were analysed for blood gas and acid-base indices at 0, 1, 2, 3, 4, 5, 6, 12, 24 and 48 h of storage. In comparison to the baseline value (0, there were significant decreases of blood pH of samples stored at RT and in the incubator after 1 h (p<0.05, the pH value of refrigerated blood samples exhibited insignificant changes during the study (p<0.05. Mean values of pCO2 showed a significant increase in Group I and Group III after 1 h then a progressive decrease after 12 h in all Groups. Mean pO2 values were significantly higher for Group I after 2 h and for Groups II and III after 1 h (p<0.05. In general, base excess decreased significantly for all the groups during the study especially in Groups II and III. In comparison with baseline values, in all groups, bicarbonate (HCO3 increased between 1 h and 6 h (p<0.05, and later decreased at the end of the study (p<0.05. In conclusion, status of acid-base indices of the samples stored at refrigerator and RT were found within normal reference range and it may be of clinical diagnostic use for up to 6 h.

  11. Photoacoustic & pulsed laser-doppler monitoring of blood concentration and perfusion in tissue

    OpenAIRE

    Kolkman, Roy Gerardus Maria

    2002-01-01

    In medicine there are situations that require continuous bedside monitoring of perfusion and blood volume. Such a situation can occur when children are born prematurely. Those children have a high risk to develop a handicap, which is most likely caused by cerebral damage due to impaired brain perfusion. Monitoring cerebral perfusion can provide information for preventive treatment. In this thesis we have explored the potential of photoacoustics and laser Doppler flowmetry for measurement of c...

  12. Determination of natural in vivo noble-gas concentrations in human blood.

    Directory of Open Access Journals (Sweden)

    Yama Tomonaga

    Full Text Available Although the naturally occurring atmospheric noble gases He, Ne, Ar, Kr, and Xe possess great potential as tracers for studying gas exchange in living beings, no direct analytical technique exists for simultaneously determining the absolute concentrations of these noble gases in body fluids in vivo. In this study, using human blood as an example, the absolute concentrations of all stable atmospheric noble gases were measured simultaneously by combining and adapting two analytical methods recently developed for geochemical research purposes. The partition coefficients determined between blood and air, and between blood plasma and red blood cells, agree with values from the literature. While the noble-gas concentrations in the plasma agree rather well with the expected solubility equilibrium concentrations for air-saturated water, the red blood cells are characterized by a distinct supersaturation pattern, in which the gas excess increases in proportion to the atomic mass of the noble-gas species, indicating adsorption on to the red blood cells. This study shows that the absolute concentrations of noble gases in body fluids can be easily measured using geochemical techniques that rely only on standard materials and equipment, and for which the underlying concepts are already well established in the field of noble-gas geochemistry.

  13. Improvement of a sensor unit for wrist blood pressure monitoring system

    Science.gov (United States)

    Koo, Sangjun; Kwon, Jongwon; Park, Yongman; Ayuzenara, Odgerel; Kim, Hiesik

    2007-12-01

    A blood pressure sensor unit for ubiquitous healthcare monitoring was newly developed. The digital wrist band-type blood pressure devices for home are popular already in the market. It is useful for checking blood pressure level at home and control of hypertension. Especially, it is very essential home device to check the health condition of blood circulation disease. Nowadays many product types are available. But the measurement of blood pressure is not accurate enough compared with the mechanical type. It needs to be upgraded to assure the precise health data enough to use in the hospital. The structure, feature and output signal of capacitor type pressure sensors are analyzed. An improved design of capacitor sensor is suggested. It shows more precise health data after use on a wrist band type health unit. They can be applied for remote u-health medical service.

  14. Tissue-Informative Mechanism for Wearable Non-invasive Continuous Blood Pressure Monitoring

    Science.gov (United States)

    Woo, Sung Hun; Choi, Yun Young; Kim, Dae Jung; Bien, Franklin; Kim, Jae Joon

    2014-10-01

    Accurate continuous direct measurement of the blood pressure is currently available thru direct invasive methods via intravascular needles, and is mostly limited to use during surgical procedures or in the intensive care unit (ICU). Non-invasive methods that are mostly based on auscultation or cuff oscillometric principles do provide relatively accurate measurement of blood pressure. However, they mostly involve physical inconveniences such as pressure or stress on the human body. Here, we introduce a new non-invasive mechanism of tissue-informative measurement, where an experimental phenomenon called subcutaneous tissue pressure equilibrium is revealed and related for application in detection of absolute blood pressure. A prototype was experimentally verified to provide an absolute blood pressure measurement by wearing a watch-type measurement module that does not cause any discomfort. This work is supposed to contribute remarkably to the advancement of continuous non-invasive mobile devices for 24-7 daily-life ambulatory blood-pressure monitoring.

  15. A novel automated discontinuous venous blood monitoring system for ex vivo glucose determination in humans.

    Science.gov (United States)

    Schaller, R; Feichtner, F; Köhler, H; Bodenlenz, M; Plank, J; Wutte, A; Mader, J K; Ellmerer, M; Hellmich, R; Wedig, H; Hainisch, R; Pieber, T R; Schaupp, L

    2009-03-15

    Intensive insulin therapy reduces mortality and morbidity in critically ill patients but imposes great demands on medical staff who must take frequent blood samples for the determination of glucose levels. A solution to this resourcing problem would be provided by an automated blood monitoring system. The aim of the present clinical study was to evaluate such a system comprising an automatic blood sampling unit linked to a glucose biosensor. Our approach was to determine the correlation and system error of the sampling unit alone and of the combined system with respect to reference levels over 12h in humans. Two venous cannulae were inserted to connect the automatic and reference systems to the subjects. Blood samples were taken at 15 and 30 min intervals. The median Pearson coefficient of correlation between manually and automatically withdrawn blood samples was 0.982 for the sampling unit alone and 0.950 for the complete system. The biosensor had a linear range up to 20 mmoll(-1) and a 95% response time of Titration Error Grid analysis suggested an acceptable treatment in 99.56% of cases. Implementation of a "Keep Vein Open" saline infusion into the automated blood sampling system reduced blood withdrawal failures through occluded catheters fourfold. In summary, automated blood sampling from a peripheral vein coupled with automatic glucose determination is a promising alternative to frequent manual blood sampling. PMID:19135351

  16. Development of portable health monitoring system for automatic self-blood glucose measurement

    Science.gov (United States)

    Kim, Huijun; Mizuno, Yoshihumi; Nakamachi, Eiji; Morita, Yusuke

    2010-02-01

    In this study, a new HMS (Health Monitoring System) device is developed for diabetic patient. This device mainly consists of I) 3D blood vessel searching unit and II) automatic blood glucose measurement (ABGM) unit. This device has features such as 1)3D blood vessel location search 2) laptop type, 3) puncturing a blood vessel by using a minimally invasive micro-needle, 4) very little blood sampling (10μl), and 5) automatic blood extraction and blood glucose measurement. In this study, ABGM unit is described in detail. It employs a syringe type's blood extraction mechanism because of its high accuracy. And it consists of the syringe component and the driving component. The syringe component consists of a syringe itself, a piston, a magnet, a ratchet and a micro-needle whose inner diameter is about 80μm. And the syringe component is disposable. The driving component consists of body parts, a linear stepping motor, a glucose enzyme sensor and a slider for accurate positioning control. The driving component has the all-in-one mechanism with a glucose enzyme sensor for compact size and stable blood transfer. On designing, required thrust force to drive the slider is designed to be greater than the value of the blood extraction force. Further, only one linear stepping motor is employed for blood extraction and transportation processes. The experimental result showed more than 80% of volume ratio under the piston speed 2.4mm/s. Further, the blood glucose was measured successfully by using the prototype unit. Finally, the availability of our ABGM unit was confirmed.

  17. 21 CFR 874.3950 - Transcutaneous air conduction hearing aid system.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Transcutaneous air conduction hearing aid system... Transcutaneous air conduction hearing aid system. (a) Identification. A transcutaneous air conduction hearing aid... occluding the ear canal. The device consists of an air conduction hearing aid attached to a...

  18. Patients’ blood pressure knowledge, perceptions and monitoring practices in community pharmacies

    Directory of Open Access Journals (Sweden)

    Lam JY

    2010-09-01

    Full Text Available Hypertension is a modifiable risk factor for cardiovascular disease. Despite this, patients often cannot or inaccurately estimate their risk factors.Objectives: In order to improve pharmacist interventions, we sought to: 1 find out patients’ knowledge about blood pressure (BP and their self- monitoring behaviors and 2 identify the relationships between these two elements. Specifically, if evaluation of BP control were related to knowledge of one’s BP level and self-monitoring habits, and if knowledge of one’s target and BP level varied with monitoring habits. Methods: Final year pharmacy students were trained and interviewed patients in community pharmacies as a required exercise in their pharmacy clerkship. Each student recruited a convenience sample of 5-10 patients who were on hypertension medication, and surveyed them regarding their BP targets, recent BP levels as well as monthly and home BP monitoring practices. Results: One third of the 449 patients interviewed were able to report a blood pressure target with 26% reporting a JNC 7 recognized target. Three quarters of patients who reported a blood pressure target were able to report a blood pressure level, with 12% being at their self- reported target. Roughly two thirds of patients perceived their BP to be “about right”, and slightly less than a third thought it to be “high”. Sixty percent of patients monitor their BP monthly, but less than 50% of patients practice home BP monitoring. Conclusions: This study along with others before it point to the knowledge and self-management gaps in patients with chronic conditions. Furthermore, pharmacy students were able to use a brief intervention to screen patients during routine care. Pharmacists can help improve patient understanding and promote increased self-management through regular BP monitoring.

  19. Ambulatory blood pressure monitoring of patients with heart failure: a new prognosis marker

    Directory of Open Access Journals (Sweden)

    Manoel F. Canesin

    2002-01-01

    Full Text Available OBJECTIVE: To evaluate the relationship between 24-hour ambulatory arterial blood pressure monitoring and the prognosis of patients with advanced congestive heart failure. METHODS: We studied 38 patients with NYHA functional class IV congestive heart failure, and analyzed left ventricular ejection fraction, diastolic diameter, and ambulatory blood pressure monitoring data. RESULTS: Twelve deaths occurred. Left ventricular ejection fraction (35.2±7.3% and diastolic diameter (72.2±7.8mm were not correlated with the survival. The mean 24-hour (SBP24, waking (SBPw, and sleeping (SBPs systolic pressures of the living patients were higher than those of the deceased patients and were significant for predicting survival. Patients with mean SBP24, SBPv, and SBPs > or = 105mmHg had longer survival (p=0.002, p=0.01 and p=0.0007, respectively. Patients with diastolic blood pressure sleep decrements (dip and patients with mean blood pressure dip or = 105 mmHg CONCLUSION: Ambulatory blood pressure monitoring appears to be a useful method for evaluating patients with congestive heart failure.

  20. Polarization Sensitive Optical Coherence Tomography for Blood Glucose Monitoring in Human Subjects

    CERN Document Server

    Solanki, Jitendra; Sen, Pratima; Andrews, Joseph Thomas

    2012-01-01

    A device based on Polarization sensitive optical coherence tomography is developed to monitor blood glucose levels in human subjects. The device was initially tested with tissue phantom. The measurements with human subjects for various glucose concentration levels are found to be linearly dependent on the degree of circular polarization obtainable from the PS-OCT.

  1. Electronic monitoring of adherence, treatment of hypertension, and blood pressure control

    NARCIS (Netherlands)

    Onzenoort, H.A. van; Verberk, W.J.; Kroon, A.A.; Kessels, A.G.; Neef, C.; Kuy, P.H. van der; Leeuw, P.W. de

    2012-01-01

    BACKGROUND: Although it is generally acknowledged that electronic monitoring of adherence to treatment improves blood pressure (BP) control by increasing patients' awareness to their treatment, little information is available on the long-term effect of this intervention. METHODS: In this observation

  2. Monitoring of tissue optical properties using OCT: application for blood glucose analysis

    Science.gov (United States)

    Larin, Kirill V.; Eledrisi, Mohsen S.; Ashitkov, Taras V.; Motamedi, Massoud; Esenaliev, Rinat O.

    2002-07-01

    Noninvasive monitoring of tissue optical properties in real time could significantly improve diagnostics and management of various diseases. Recently we proposed to use high- resolution Optical Coherence Tomography (OCT) technique for measurement of tissue scattering coefficient at the depth of up to 1mm. Our pilot studies performed in vitro and in vivo demonstrated that measurement of tissue scattering with this technique can potentially be applied for noninvasive monitoring of blood glucose concentration. High resolution and coherent photon detection of the OCT technique allowed detection of glucose-induced changes in the scattering coefficient. In this paper we report results of in vivo studies performed in dog, New Zealand rabbits, and first human subjects. OCT system with the wavelength of 1300 nm was used in our experiments. OCT signal slope was measured and compared with actual blood glucose concentration. Bolus glucose injections and glucose clamping administrations were used in animal studies. OCT signals were recorded form human subjects during oral glucose tolerance test. Results obtained form both animal and human studies show good correlation between slope of the OCT signals and actual blood glucose concentration measured using standard glucometesr. Sensitivity and accuracy of blood glucose concentrations monitoring with the OCT is discussed. Obtained result suggest that OCT is a promising technique for noninvasive monitoring of tissue analytes including glucose.

  3. Self-monitoring of blood glucose in tablet-treated type 2 diabetic patients (ZODIAC)

    NARCIS (Netherlands)

    Kleefstra, Nanno; Hortensius, J.; Logtenberg, Susan; Slingerland, R. J.; Groenier, K. H.; Houweling, S. T.; Gans, R. O. B.; van Ballegooie, E; Bilo, H. J. G.

    2010-01-01

    Background: Whether self-monitoring of blood glucose (SMBG) improves glycaemic control in patients with type 2 diabetes mellitus (T2DM) not using insulin is questionable. Our aim was to investigate the effects of SMBG in patients with T2DM who were in persistent moderate glycaemic control whilst not

  4. What do professionals recommend regarding the frequency of self-monitoring of blood glucose?

    NARCIS (Netherlands)

    Hortensius, J.; Kleefstra, N.; Houweling, S. T.; van der Bijl, J. J.; Gans, R. O. B.; Bilo, H. J. G.

    2012-01-01

    Background: Patients' adherence to guidelines regarding self-monitoring of blood glucose (SMBG) is limited. However, there are no previous reports about the recommendations that are given in clinical practice concerning SMBG. The aim of this study was to investigate what healthcare providers recomme

  5. A comprehensive evaluation of strip performance in multiple blood glucose monitoring systems.

    Science.gov (United States)

    Katz, Laurence B; Macleod, Kirsty; Grady, Mike; Cameron, Hilary; Pfützner, Andreas; Setford, Steven

    2015-05-01

    Accurate self-monitoring of blood glucose is a key component of effective self-management of glycemic control. Accurate self-monitoring of blood glucose results are required for optimal insulin dosing and detection of hypoglycemia. However, blood glucose monitoring systems may be susceptible to error from test strip, user, environmental and pharmacological factors. This report evaluated 5 blood glucose monitoring systems that each use Verio glucose test strips for precision, effect of hematocrit and interferences in laboratory testing, and lay user and system accuracy in clinical testing according to the guidelines in ISO15197:2013(E). Performance of OneTouch(®) VerioVue™ met or exceeded standards described in ISO15197:2013 for precision, hematocrit performance and interference testing in a laboratory setting. Performance of OneTouch(®) Verio IQ™, OneTouch(®) Verio Pro™, OneTouch(®) Verio™, OneTouch(®) VerioVue™ and Omni Pod each met or exceeded accuracy standards for user performance and system accuracy in a clinical setting set forth in ISO15197:2013(E). PMID:25702769

  6. Photoacoustic & pulsed laser-doppler monitoring of blood concentration and perfusion in tissue

    NARCIS (Netherlands)

    Kolkman, Roy Gerardus Maria

    2002-01-01

    In medicine there are situations that require continuous bedside monitoring of perfusion and blood volume. Such a situation can occur when children are born prematurely. Those children have a high risk to develop a handicap, which is most likely caused by cerebral damage due to impaired brain perfus

  7. Comparison of serum bilirubin estimation with transcutaneous bilirubinometry in neonates

    International Nuclear Information System (INIS)

    Objective: To assess usefulness of Minolta Air shield transcutaneous bilirubinometer by comparing bilirubin values obtained by transcutaneous jaundice meter with serum bilirubin estimation. Design: Analytical cross sectional study. Place and duration: NICU Military Hospital Rawalpindi Pakistan Jun 2002 to May 2005. Subjects and Methods: One hundred and fifty neonates admitted to NICU because of visible jaundice were included in the study. Serum was sent to laboratory for total bilirubin estimation. At the same time bilirubin was also checked by a Jaundice Meter. Data was tabulated and t-test applied to compare the two values. Results: One hundred and fifty paired estimations were performed. The transcutaneous bilirubin values ranged from 8.0 mg/dl to 20.4 mg/dl. While serum bilirubin by jaundice meter values ranged between 5.3 mg/dl and 26.0 mg/dl. A Scatter diagram was plotted. It showed a correlation coefficient of 0.78. Conclusion: Bilirubin values obtained by transcutaneous bilirubin meter were not significantly different from laboratory values thus proving the fact that transcutaneous bilirubinometer is a useful device to measure bilirubin. (author)

  8. The food and drug administration is now preparing to establish tighter performance requirements for blood glucose monitors.

    Science.gov (United States)

    Klonoff, David C

    2010-05-01

    On March 16 and 17, 2010, the Food and Drug Administration (FDA) presented a public meeting about blood glucose monitoring at the Gaithersberg Hilton Hotel. The meeting was intended to present expert opinions and solicit input from the public about whether to develop new regulatory policies for blood glucose monitors. The meeting was divided into three sections: (1) Clinical Accuracy Requirements for Blood Glucose Monitors, (2) Interferences and Limitations of Blood Glucose Monitors, and (3) Tight Glycemic Control. Many officials from the Center for Devices and Radiologic Health and the Office of In Vitro Diagnostic Devices, which are the parts of FDA that regulate approval of blood glucose monitors, either spoke on the agenda or attended in the audience. Approximately 300 people attended; they were mostly clinicians (such as adult endocrinologists, pediatric endocrinologists, internists, clinical chemists, intensivists, surgeons, nurses, and diabetes educators) or industry officials from companies involved in glucose monitoring, pharmaceutical products, data analysis, or regulatory consulting. PMID:20513313

  9. Ambulatory Blood Pressure Monitoring: Five Decades of More Light and Less Shadows

    Science.gov (United States)

    Nobre, Fernando; Mion Junior, Décio

    2016-01-01

    Casual blood pressure measurements have been extensively questioned over the last five decades. A significant percentage of patients have different blood pressure readings when examined in the office or outside it. For this reason, a change in the paradigm of the best manner to assess blood pressure has been observed. The method that has been most widely used is the Ambulatory Blood Pressure Monitoring - ABPM. The method allows recording blood pressure measures in 24 hours and evaluating various parameters such as mean BP, pressure loads, areas under the curve, variations between daytime and nighttime, pulse pressure variability etc. Blood pressure measurements obtained by ABPM are better correlated, for example, with the risks of hypertension. The main indications for ABPM are: suspected white coat hypertension and masked hypertension, evaluation of the efficacy of the antihypertensive therapy in 24 hours, and evaluation of symptoms. There is increasing evidence that the use of ABPM has contributed to the assessment of blood pressure behaviors, establishment of diagnoses, prognosis and the efficacy of antihypertensive therapy. There is no doubt that the study of 24-hour blood pressure behavior and its variations by ABPM has brought more light and less darkness to the field, which justifies the title of this review. PMID:27168473

  10. Carotid arterial blood pressure waveform monitoring using a portable ultrasound system.

    Science.gov (United States)

    Joohyun Seo; Pietrangelo, Sabino J; Hae-Seung Lee; Sodini, Charles G

    2015-08-01

    This work presents a non-invasive arterial blood pressure (ABP) waveform monitoring technique using ultrasound. A portable ultrasound system to excite ultrasound transducers and acquire data is designed with off-the-shelf components. The insonation angles are identified using a vector Doppler technique based on the cosine dependency of the Doppler signals. The pulse pressure of an estimated waveform at the left common carotid artery is compared to the standard sphygmomanometer measurement in a clinical test. The estimated carotid ABP waveform shows excellent agreement to the finger ABP waveform with expected discrepancy of the systolic peak shape due to different measurement sites. The proposed method also tracks slow blood pressure fluctuations. This validation on human subjects shows potential for a noninvasive blood pressure waveform monitoring device at central arterial sites. PMID:26737584

  11. Gas Analysis and Monitoring Systems for the RPC Detector of CMS at LHC

    CERN Document Server

    Abbrescia, M; Guida, R; Iaselli, G; Liuzzi, R; Loddo, F; Maggi, M; Marangelli, B; Natali, S; Nuzzo, S; Pugliese, G; Ranieri, A; Romano, F; Trentadue, R; Benussi, L; Bertani, M; Bianco, S; Caponero, M A; Colonna, D; Donisi, D; Fabbri, F L; Felli, F; Ortenzi, M G B; Pallotta, M; Paolozzi, A; Passamonti, L; Ponzio, B; Pucci, C; Polese, G S G; Segoni, I; Cavallo, N; Fabozzi, F; Paolucci, P; Piccolo, D; Belli, C S G; Grelli, A; Necchi, M; Ratti, S P; Riccardi, C; Torre, P; Vitulo, P

    2006-01-01

    The Resistive Plate Chambers (RPC) detector of the CMS experiment at the LHC proton collider (CERN, Switzerland) will employ an online gas analysis and monitoring system of the freon-based gas mixture used. We give an overview of the CMS RPC gas system, describe the project parameters and first results on gas-chromatograph analysis. Finally, we report on preliminary results for a set of monitor RPC.

  12. Circulating miRNA-155 is realted to blood pressure monitoring parameters

    Institute of Scientific and Technical Information of China (English)

    WU Ying; HUANG Yu-qing; HUANG Cheng; LI Jie; CAI An-ping; YU Xue-ju; ZHOU Dan

    2016-01-01

    Background Studies have shown that miRNA-155 played an important role in the process of development of hypertension.However,there is no date about miRNA-155 and blood pressure monitoring parameters.Therefore,we examined whether in hypertensive patients the expression level of plasma miRNA-155 related to 24-h ambulatory blood pressure monitoring (ABPM) parameters.Methods A cohort of adult patients scheduled to receive physical examination,office and ambulatory blood pressure monitoring.Quantitative reverse transcriptase polymerase chain reaction (qRT-PCR) was used to evaluate the expression of selected miRNA-155.The miRNA-155 expression level correlation between blood pressure parameters was assessed using the Spearman correlation coefficient.Results Fifty four essential hypertension patients (25 men;mean age,53.28 ± 9.52 years) and thirty healthy volunteers (15 men;mean age,53.03 ± 5.87 years) were included.We observed higher expression level of miRNA-155 (32.31 ± 2.85 vs 27.21 ± 1.59,P < 0.001) in hypertensive patients compared to healthy control individuals.MiRNA-155 expression level showed significant positive correlation with 24 h Daytime SBP (r =0.681,P < 0.001),24 h Daytime DBP (r =0.473,P < 0.001),24 h Daytime PP (r =0.565,P < 0.001) and dipping (r =0.257,P =0.018),respectively.Conclusions Our study showed that miRNA-155 expression level was associated positively with daytime blood pressure monitoring parameters,as well as blood pressure variability,indicating a possible implication of miRNA-155 in the pathogenesis of hypertension.

  13. Upgrade to the Cryogenic Hydrogen Gas Target Monitoring System

    Science.gov (United States)

    Slater, Michael; Tribble, Robert

    2013-10-01

    The cryogenic hydrogen gas target at Texas A&M is a vital component for creating a secondary radioactive beam that is then used in experiments in the Momentum Achromat Recoil Spectrometer (MARS). A stable beam from the K500 superconducting cyclotron enters the gas cell and some incident particles are transmuted by a nuclear reaction into a radioactive beam, which are separated from the primary beam and used in MARS experiments. The pressure in the target chamber is monitored so that a predictable isotope production rate can be assured. A ``black box'' received the analog pressure data and sent RS232 serial data through an outdated serial connection to an outdated Visual Basic 6 (VB6) program, which plotted the chamber pressure continuously. The black box has been upgraded to an Arduino UNO microcontroller [Atmel Inc.], which can receive the pressure data and output via USB to a computer. It has been programmed to also accept temperature data for future upgrade. A new computer program, with updated capabilities, has been written in Python. The software can send email alerts, create audible alarms through the Arduino, and plot pressure and temperature. The program has been designed to better fit the needs of the users. Funded by DOE and NSF-REU Program.

  14. Noble Gas Measurement and Analysis Technique for Monitoring Reprocessing Facilities

    Energy Technology Data Exchange (ETDEWEB)

    Charlton, William S

    1999-09-01

    An environmental monitoring technique using analysis of stable noble gas isotopic ratios on-stack at a reprocessing facility was developed. This technique integrates existing technologies to strengthen safeguards at reprocessing facilities. The isotopic ratios are measured using a mass spectrometry system and are compared to a database of calculated isotopic ratios using a Bayesian data analysis method to determine specific fuel parameters (e.g., burnup, fuel type, fuel age, etc.). These inferred parameters can be used by investigators to verify operator declarations. A user-friendly software application (named NOVA) was developed for the application of this technique. NOVA included a Visual Basic user interface coupling a Bayesian data analysis procedure to a reactor physics database (calculated using the Monteburns 3.01 code system). The integrated system (mass spectrometry, reactor modeling, and data analysis) was validated using on-stack measurements during the reprocessing of target fuel from a U.S. production reactor and gas samples from the processing of EBR-II fast breeder reactor driver fuel. These measurements led to an inferred burnup that matched the declared burnup with sufficient accuracy and consistency for most safeguards applications. The NOVA code was also tested using numerous light water reactor measurements from the literature. NOVA was capable of accurately determining spent fuel type, burnup, and fuel age for these experimental results. Work should continue to demonstrate the robustness of this system for production, power, and research reactor fuels.

  15. Noble Gas Measurement and Analysis Technique for Monitoring Reprocessing Facilities

    International Nuclear Information System (INIS)

    An environmental monitoring technique using analysis of stable noble gas isotopic ratios on-stack at a reprocessing facility was developed. This technique integrates existing technologies to strengthen safeguards at reprocessing facilities. The isotopic ratios are measured using a mass spectrometry system and are compared to a database of calculated isotopic ratios using a Bayesian data analysis method to determine specific fuel parameters (e.g., burnup, fuel type, fuel age, etc.). These inferred parameters can be used by investigators to verify operator declarations. A user-friendly software application (named NOVA) was developed for the application of this technique. NOVA included a Visual Basic user interface coupling a Bayesian data analysis procedure to a reactor physics database (calculated using the Monteburns 3.01 code system). The integrated system (mass spectrometry, reactor modeling, and data analysis) was validated using on-stack measurements during the reprocessing of target fuel from a U.S. production reactor and gas samples from the processing of EBR-II fast breeder reactor driver fuel. These measurements led to an inferred burnup that matched the declared burnup with sufficient accuracy and consistency for most safeguards applications. The NOVA code was also tested using numerous light water reactor measurements from the literature. NOVA was capable of accurately determining spent fuel type, burnup, and fuel age for these experimental results. Work should continue to demonstrate the robustness of this system for production, power, and research reactor fuels

  16. Development of bremsstrahlung detection type tritium gas monitoring system, (1)

    International Nuclear Information System (INIS)

    A tritium monitoring system by means of bremsstrahlung detecting was developed. A prototype system consisted of a sampling cylinder, a gas circulating apparatus, an NaI(T1) detector, an amplifier and a multichannel analyzer. The sizes of sampling cylinders used 208 mm phi x 290; 170; 70 mmH, 133 mm phi x 292; 172; 72 mmH and 55 mm phi x 294; 174; 74 mmH, respectively. The sensitivity of prototype system was from 12 to 57 cps/μCi.cm-3, depending on the size of sampling cylinder and an efficiency of NaI(T1) detector. When pulses due to breamsstrahlungs with energy from 4 to 17 keV were counted, the minimum detectable concentration of the prototype tritium monitoring system was obtained to be 5.2 x 10-3 μCi/cm3. It was evaluated that the detectable range of concentration was from 1 x 10-2 to 1 x 103 μCi/cm3. (author)

  17. Optical methods for monitoring harmful gas in animal facilities

    Science.gov (United States)

    Zhang, Shirui; Dong, Daming; Zheng, Wengang; Wang, Jihua

    2014-06-01

    Animal facilities produce large amounts of harmful gases such as ammonia, hydrogen sulfide, and methane, many of which have a pungent odor. The harmful gases produced by animal housing not only affect the health of people and livestock but also pollute the air. The detection of the harmful gases can effectively improve efficiency of livestock production and reduce environmental pollution. More and more optical detection methods are applied to the detection of the harmful gases produced by animal housing. This summarizes optical detection methods for monitoring the harmful gases in animal housing recently, including nondispersive infrared gas analyzer, ultraviolet differential optical absorption spectroscopy, Fourier transform infrared spectroscopy, and tunable diode laser absorption spectroscopy. The basic principle and the characteristics of these methods are illustrated and the applications on the detection of harmful gases in animal housing are described. Meanwhile, the research of harmful gases monitoring for livestock production based on these methods were listed. The current situation and future development of the detection methods for harmful gases generated by animal housing were summarized by comparing the advantages and disadvantages of each method.

  18. Transcutaneous mechanical nerve stimulation using perineal vibration: a novel method for the treatment of female stress urinary incontinence

    DEFF Research Database (Denmark)

    Sønksen, Jens; Ohl, Dana A; Bonde, Birthe;

    2007-01-01

    We defined basic guidelines for transcutaneous mechanical nerve stimulation in modifying pelvic floor responses in women and determined the efficacy of transcutaneous mechanical nerve stimulation in treating stress urinary incontinence.......We defined basic guidelines for transcutaneous mechanical nerve stimulation in modifying pelvic floor responses in women and determined the efficacy of transcutaneous mechanical nerve stimulation in treating stress urinary incontinence....

  19. Cost estimation of hypertension management based on home blood pressure monitoring alone or combined office and ambulatory blood pressure measurements.

    Science.gov (United States)

    Boubouchairopoulou, Nadia; Karpettas, Nikos; Athanasakis, Kostas; Kollias, Anastasios; Protogerou, Athanase D; Achimastos, Apostolos; Stergiou, George S

    2014-10-01

    This study aims at estimating the resources consumed and subsequent costs for hypertension management, using home blood pressure (BP) monitoring (HBPM) alone versus combined clinic measurements and ambulatory blood pressure monitoring (C/ABPM). One hundred sixteen untreated hypertensive subjects were randomized to use HBPM or C/ABPM for antihypertensive treatment initiation and titration. Health resources utilized within 12-months follow-up, their respective costs, and hypertension control were assessed. The total cost of the first year of hypertension management was lower in HBPM than C/ABPM arm (€1336.0 vs. €1473.5 per subject, respectively; P cost was identical in both arms. There was no difference in achieved BP control and drug expenditure (HBPM: €233.1 per subject; C/ABPM: €247.6 per subject; P = not significant), whereas the cost of BP measurements and/or visits was higher in C/ABPM arm (€393.9 vs. €516.9, per patient, respectively P cost for subsequent years (>1) was €348.9 and €440.2 per subject, respectively for HBPM and C/ABPM arm and €2731.4 versus €3234.3 per subject, respectively (P cost than C/ABPM, and the same trend is observed in 5-year projection. The results on the resources consumption can be used to make cost estimates for other health-care systems.

  20. Analysis of mean transcutaneous capnography in consecutive patients undergoing polysomnography

    Directory of Open Access Journals (Sweden)

    Giulio Cesare Pinnola

    2014-11-01

    Full Text Available Transcutaneous capnography is a noninvasive method useful for analysis of the behavioral tendency of transcutaneous CO2 pressure (PtcCO2 in patients undergoing polysomnography, to evaluate respiratory sleep disorders. Objective Determine normative PtcCO2 values in normal patients undergoing polysomnography. Method One hundred seventy-nine patients who underwent polysomnography with simultaneous PtcCO2 measurement were assessed by means of a transcutaneous capnograph (TCM4 series from Radiomiter. Results The group classified as normal (N=53 presented a apnea/hypopnea index (AHI <5 events/per hour of sleep and their age groups varied between 7 and 76 years of age. Conclusion Global mean values of PtcCO2 in the normal group had a Gaussian distribution that varied between 33.1 and 50.0 mmHg (SD 4,363. Such findings allowed the establishment of normative PtcCO2 values for normal individuals.

  1. Transcutaneous oximetry: normal values for the lower limb.

    Science.gov (United States)

    2016-03-01

    Consistent with the Committee on Publication Ethics guidelines, we the above authors are initiating the retraction of our paper: Blake DF, Young DA, Brown LH: Transcutaneous oximetry: normal values for the lower limb. Diving Hyperb Med. 2014 September;44(3):146-153. We wish to make the following statement: "The authors voluntarily retract this article after discovering a critical error associated with the instrumentation used in the study, namely the fitting of incorrect sensor membranes on the electrodes of the transcutaneous oximetry device used in the study. This resulted in transcutaneous oxygen tension (PtcO₂) measurements that were consistently lower than those that would be recorded with the correct electrode membranes in place, which we recently confirmed by comparing the two membrane types once we discovered the error. We are in the process of replicating our work using the correct PtcO₂ specific membranes." PMID:27044466

  2. Transcutaneous oximetry measurement: normal values for the upper limb.

    Science.gov (United States)

    2016-03-01

    Consistent with the Committee on Publication Ethics guidelines, we the above authors are initiating the retraction of our paper: Young DA, Blake DF, Brown LH: Transcutaneous oximetry measurement: normal values for the upper limb. Diving Hyperb Med. 2012;42(4):208-213. We wish to make the following statement: "The authors voluntarily retract this article after discovering a critical error associated with the instrumentation used in the study, namely the fitting of incorrect sensor membranes on the electrodes of the transcutaneous oximetry device used in the study. This resulted in transcutaneous oxygen tension (PtcO₂) measurements that were consistently lower than those that would have been recorded with the correct electrode membranes in place. We recently confirmed this by comparing the two membrane types once we discovered the error. We are in the process of replicating our work using the correct PtcO₂ specific membranes." PMID:27044465

  3. Quality assessment of patients’ self-monitoring of blood glucose in community pharmacies

    Directory of Open Access Journals (Sweden)

    Kjome RL

    2010-03-01

    Full Text Available Objective: To evaluate diabetes patients’ self-monitoring of blood glucose using a community pharmacy-based quality assurance procedure, to investigate whether the procedure improved the quality of the patient performance of self monitoring of blood glucose, and to examine the opinions of the patients taking part in the study. Methods: The results of patient blood glucose measurements were compared to the results obtained with HemoCue Glucose 201+ by pharmacy employees in 16 Norwegian community pharmacies. Patient performance was monitored using an eight item checklist. Patients whose blood glucose measurements differed from pharmacy measurements by more than 20% were instructed in the correct use of their glucometer. The patients then re-measured their blood glucose. If the results were still outside the set limits, the control procedure was repeated with a new lot of glucometer strips, and then with a new glucometer. The patients returned for a follow-up visit after three months. Results: During the first visit, 5% of the 338 patients had measurements that deviated from pharmacy blood glucose values by more than 20% and user errors were observed for 50% of the patients. At the second visit, there was no significant change in the analytical quality of patient measurements, but the percentage of patients who made user errors had decreased to 29% (p < 0.001. Eighty-five percent of the patients reported that they used their blood glucose results to adjust medication, exercise or meals. Fifty-one percent of the patients reported a greater trust in their measurements after the second visit. Eighty percent of patients wished to have their measurements assessed yearly. Of these patients, 83% preferred to have the assessment done at the community pharmacy. Conclusion: A community pharmacy-based quality assessment procedure of patients’ self monitoring of blood glucose significantly reduced the number of user errors. The analytical quality of the

  4. Biomarkers for monitoring pre-analytical quality variation of mRNA in blood samples.

    Directory of Open Access Journals (Sweden)

    Hui Zhang

    Full Text Available There is an increasing need for proper quality control tools in the pre-analytical phase of the molecular diagnostic workflow. The aim of the present study was to identify biomarkers for monitoring pre-analytical mRNA quality variations in two different types of blood collection tubes, K2EDTA (EDTA tubes and PAXgene Blood RNA Tubes (PAXgene tubes. These tubes are extensively used both in the diagnostic setting as well as for research biobank samples. Blood specimens collected in the two different blood collection tubes were stored for varying times at different temperatures, and microarray analysis was performed on resultant extracted RNA. A large set of potential mRNA quality biomarkers for monitoring post-phlebotomy gene expression changes and mRNA degradation in blood was identified. qPCR assays for the potential biomarkers and a set of relevant reference genes were generated and used to pre-validate a sub-set of the selected biomarkers. The assay precision of the potential qPCR based biomarkers was determined, and a final validation of the selected quality biomarkers using the developed qPCR assays and blood samples from 60 healthy additional subjects was performed. In total, four mRNA quality biomarkers (USP32, LMNA, FOSB, TNRFSF10C were successfully validated. We suggest here the use of these blood mRNA quality biomarkers for validating an experimental pre-analytical workflow. These biomarkers were further evaluated in the 2nd ring trial of the SPIDIA-RNA Program which demonstrated that these biomarkers can be used as quality control tools for mRNA analyses from blood samples.

  5. Trans-Cutaneous Bilirubinometery versus Serum Bilirubin in Neonatal Jaundice.

    Science.gov (United States)

    Mahram, Manoochehr; Oveisi, Sonia; Jaberi, Najmeh

    2015-12-01

    Hyperbilirubinemia is a common problem in neonates and causes serious complications. Thus, serial measurements of bilirubin should be done. This assessment is done through two methods of laboratory measurement in serum sample and transcutaneous bilirubinometer. This descriptive study compared transcutaneous bilirubin assessment and laboratory serum bilirubin. Bilirubin level was assessed among 256 neonates admitted to the Qods Children's Hospital in Qazvin- Iran, because of neonatal indirect jaundice, through two methods of transcutaneous bilirubinometery from two sites of forehead and sternum and laboratory measurement of bilirubin in serum. The cases were non-hemolytic icteric term neonates weighing 2500 gram or more and had not received phototherapy or other treatments. Neonates with hemolytic forms of jaundice, sepsis and suspicious to metabolic disorders were excluded. Assessments by means of KJ-8000 transcutaneous bilirubinometer from two sites of forehead and sternum and through laboratory measurement of serum bilirubin were registered and analyzed. The results of the current study showed that there was a correlation of 0.82 between serum bilirubin and transcutaneous forehead bilirubin assessment and for the used device sensitivity of 0.844; specificity of 0.842, Youden Index of 0.709 and Shortest of 0.042 for a cut-off of 12.4 in bilirubin of participants. Furthermore, Likelihood Ratio positive and negative (LR) were 5.665 and 0.164, respectively and diagnostic Odds Ratio (LR+/LR-) was 34.56. Transcutaneous bilirubinometery can be considered as a reliable tool to assess bilirubin for the screening of neonatal jaundice in term neonates. PMID:26749233

  6. Trans-Cutaneous Bilirubinometery versus Serum Bilirubin in Neonatal Jaundice

    Directory of Open Access Journals (Sweden)

    Manoochehr Mahram

    2015-12-01

    Full Text Available Hyperbilirubinemia is a common problem in neonates and causes serious complications. Thus, serial measurements of bilirubin should be done. This assessment is done through two methods of laboratory measurement in serum sample and transcutaneous bilirubinometer. This descriptive study compared transcutaneous bilirubin assessment and laboratory serum bilirubin. Bilirubin level was assessed among 256 neonates admitted to the Qods Children’s Hospital in Qazvin- Iran, because of neonatal indirect jaundice, through two methods of transcutaneous bilirubinometery from two sites of forehead and sternum and laboratory measurement of bilirubin in serum. The cases were non-hemolytic icteric term neonates weighing 2500 gram or more and had not received phototherapy or other treatments. Neonates with hemolytic forms of jaundice, sepsis and suspicious to metabolic disorders were excluded. Assessments by means of KJ-8000 transcutaneous bilirubinometer from two sites of forehead and sternum and through laboratory measurement of serum bilirubin were registered and analyzed. The results of the current study showed that there was a correlation of 0.82 between serum bilirubin and transcutaneous forehead bilirubin assessment and for the used device sensitivity of 0.844; specificity of 0.842, Youden Index of 0.709 and Shortest of 0.042 for a cut-off of 12.4 in bilirubin of participants. Furthermore, Likelihood Ratio positive and negative (LR were 5.665 and 0.164, respectively and diagnostic Odds Ratio (LR+/LR- was 34.56. Transcutaneous bilirubinometery can be considered as a reliable tool to assess bilirubin for the screening of neonatal jaundice in term neonates.

  7. Correlation between arterial and venous blood gas analysis parameters in patients with acute exacerbation of chronic obstructive pulmonary disease

    Directory of Open Access Journals (Sweden)

    Novović Miloš

    2012-01-01

    Full Text Available Introduction. Arterial blood gas (ABG analyses have an important role in the assessment and monitoring of the metabolic and oxygen status of patients with acute exacerbation of chronic obstructive pulmonary disease (COPD. Arterial puncture could have a lot of adverse effects, while sampling of venous blood is simpler and is not so invasive. Objective. The aim of this study was to evaluate whether venous blood gas (VBG values of pH, partial pressure of carbon dioxide (PCO2, partial oxygen pressure (PO2, bicarbonate (HCO3, and venous and arterial blood oxygen saturation (SO2 can reliably predict ABG levels in patients with acute exacerbation of COPD. Methods. Forty-seven patients with a prior diagnosis of COPD were included in this prospective study. The patients with acute exacerbation of this disease were examined at the General Hospital EMS Department in Prijepolje. ABG samples were taken immediately after venous sampling, and both were analyzed. Results. The Pearson correlation coefficients between arterial and venous parameters were 0.828, 0.877, 0.599, 0.896 and 0.312 for pH, PCO2, PO2, HCO3 and SO2, respectively. The statistically significant correlation between arterial and venous pH, PCO2 and HCO3, values was found in patients with acute exacerbation of COPD (p<0.001. Conclusion. When we cannot provide arterial blood for analysis, venous values of the pH, Pv,CO2 and HCO3 parameters can be an alternative to their arterial equivalents in the interpretation of the metabolic status in patients with acute exacerbation of COPD, while the values of venous Pv,O2 and Sv,O2 cannot be used as predictors in the assessment of oxygen status of such patients.

  8. [Influence of treatment with olmesartan on ambulatory blood pressure monitoring parameters in patients with arterial hypertension].

    Science.gov (United States)

    Bregvadze, T R; Tseluĭko, V I; Mishchuk, N E

    2013-12-01

    Hypertension is the most common disease of the cardiovascular system. Active treatment of hypertension with adequate control of blood pressure (BP) can prevent complications, improve life quality and increase life expectancy. One of the interesting new antihypertensive agents, from the group of angiotensin receptor blockers is olmesartan. The obvious advantages of ambulatory blood pressure monitoring to traditional one-time measurements of BP make this method perspective for quality control of anti-hypertensive therapy. The aim of this study was to evaluate the influence of treatment with olmesartan on ambulatory blood pressure monitoring parameters in patients with hypertension. 38 out-patients with hypertension at the age of 25-84 years (mean 55,3±10,6) were studied. Patients received olmesartan 20 mg daily as monotherapy (20 patients (52,6%)) or in combination with other antihypertensive agents (18 patients (47,4%)). Treatment continued for 6 months. The complex examination included: measurement of office brachial BP, electrocardiography, echocardiography and ambulatory blood pressure monitoring (ABPM). As a result of treatment, office BP and diurnal BP, according to ABPM, significantly decreased; the favorable circadian BP profile dynamics were found: significantly less frequently observed lack of reduction in BP during night (daily index - non-dipper) - 18% vs. 64% (p treatment of hypertensive patients with olmesartan provides significant decline not only in office BP, but also in diurnal BP, normalizes BP of active and passive periods, also - daily index and reduces BP variability.

  9. Systemic and mucosal immune response induced by transcutaneous immunization using Hepatitis B surface antigen-loaded modified liposomes.

    Science.gov (United States)

    Mishra, Dinesh; Mishra, Pradyumna Kumar; Dubey, Vaibhav; Nahar, Manoj; Dabadghao, Sunil; Jain, N K

    2008-04-23

    We have evaluated the efficiency of novel modified liposomes (ethosomes) for transcutaneous immunization (TCI) against Hepatitis B. Antigen-loaded ethosomes were prepared and characterized for shape, lamellarity, fluidity, size distribution, and entrapment efficiency. Spectral bio-imaging and flow cytometric studies showed efficient uptake of Hepatitis B surface antigen (HBsAg)-loaded ethosomes by murine dendritic cells (DCs) in vitro, reaching a peak by 180 min. Transcutaneous delivery potential of the antigen-loaded system using human cadaver skin demonstrated a much higher skin permeation of the antigen in comparison to conventional liposomes and soluble antigen preparation. Topically applied HBsAg-loaded ethosomes in experimental mice showed a robust systemic and mucosal humoral immune response compared to intramuscularly administered alum-adsorbed HBsAg suspension, topically applied plain HBsAg solution and hydroethanolic (25%) HBsAg solution. The ability of the antigen-pulsed DCs to stimulate autologous peripheral blood lymphocytes was demonstrated by BrdU assay and a predominantly TH1 type of immune response was observed by multiplex cytometric bead array analysis. HBsAg-loaded ethosomes are able to generate a protective immune response and their ability to traverse and target the immunological milieu of the skin may find a potential application in the development of a transcutaneous vaccine against Hepatitis B virus (HBV).

  10. Validation protocol for multiple blood gas analyzers in accordance with laboratory accreditation programs

    OpenAIRE

    Pérsio A. R. Ebner; Paschoalina Romano; Alexandre Sant’Anna; Maria Elizabete Mendes; Magna Oliveira; Nairo M. Sumita

    2015-01-01

    ABSTRACTIntroduction:The results of blood gas analysis using different instrumentation can vary widely due to the methodological differences, the calibration procedures and the use of different configurations for each type of instrument.Objective:The objective of this study was to evaluate multiple analytical systems for measurement of blood gases, electrolytes and metabolites in accordance with the accreditation program (PALC) of Sociedade Brasileira de Patologia Clínica/Medicina Laboratoria...

  11. Patient self-monitoring of blood glucose and refinements of conventional insulin treatment.

    Science.gov (United States)

    Tattersall, R; Gale, E

    1981-01-01

    The compelling evidence that blood glucose control will slow or prevent microvascular complications has stimulated research to find better ways of managing insulin-dependent diabetes. The excellent results obtained with "open loop" insulin infusion systems suggest that the relative failure of conventional treatment is the result of (1) a lack of appropriate feedback to the patient and (2) the use of insulin regimens which do not mimic physiologic insulinemia, particularly in the basal state. Doctors regard blood glucose measurements as an essential part of diabetic management and extension of this technology to patients has added a new dimension, particularly in the assessment of control. Nevertheless, home blood-glucose monitoring will not necessarily improve diabetic control; the best results have been obtained when it has been offered as part of a package deal which includes more investment of time and interest by patients and doctor together with joint discussions of problems and changes in treatment. The biggest problem with conventional twice daily insulin regimens is to sustain constant basal insulin levels during the night. Attempts to obtain fasting normoglycemia with an injection before supper often result in nocturnal hyperinsulinemia and hypoglycemia. This can usually be resolved by changing to a three times daily regimen with an extra injection of NPH insulin at bedtime. Three times daily insulin injections with feedback from home blood-glucose monitoring give as good blood glucose control as infusion systems and are cheaper and more acceptable to patients. PMID:7006390

  12. Mathematical modeling of blood-gas kinetics for the volatile organic compounds isoprene and acetone

    International Nuclear Information System (INIS)

    Breath gas analysis is based on the compelling concept that the exhaled breath levels of endogenously produced volatile organic compounds (VOCs) can provide a direct, non-invasive window to the blood and hence, by inference, to the body. In this sense, breath VOCs are regarded as a comprehensive repository of valuable physiological and clinical information, that might be exploited in such diverse areas as diagnostics, therapeutic monitoring or general dynamic assessments of metabolic function, pharmacodynamics (e.g., in drug testing) and environmental exposure (e.g., in occupational health). Despite this enormous potential, the lack of standardized breath sampling regimes as well as the poor mechanistic understanding of VOC exhalation kinetics could cast a cloud over the widespread use of breath gas analysis in the biomedical sciences. In this context, a primary goal of the present thesis is to provide a better quantitative insight into the breath behavior of two prototypic VOCs, isoprene and acetone. A compartmental modeling framework is developed and validated by virtue of real-time breath measurements of these trace gases during distinct physiological states. In particular, the influence of various hemodynamic and ventilatory parameters on VOC concentrations in exhaled breath is investigated. This approach also complements previous steady state investigations in toxicology. From a phenomenological point of view, both acetone and isoprene concentrations in end-tidal breath are demonstrated to exhibit a reproducible non-steady state behavior during moderate workload challenges on a stationary bicycle. However, these dynamics depart drastically from what is expected on the basis of classical pulmonary inert gas elimination theory. More specifically, the start of exercise is accompanied by an abrupt increase in breath isoprene levels, usually by a factor of 3 to 4 compared with the steady state value during rest. This phase is followed by a gradual decline and the

  13. Effects of Imidapril on Venous Blood Gas Values in Broiler Chickens Exposed to Low Ambient Temperature

    Institute of Scientific and Technical Information of China (English)

    Xueqin HAO; Meng LI; Shouyan ZHANG; Yongshu WANG; Tongwen SUN

    2013-01-01

    [Objective] This study was designed to evaluate the effects of imidapril on blood gas parameters in broiler chickens.[Method] Twenty-four chickens were randomly divided into three groups (n=8),control group,low temperature group and imidapril group.Chickens in low temperature group and imidapril group were exposed to low ambient temperature (12-18 ℃) from age at 14 d to 45 d,whereas the control group was exposed to 24-30 ℃; chickens in imidapril group were gavaged with imidapril (3 mg/kg) once daily for 30 d.At age of 45 d,blood was taken from wing vein and blood gas parameters were evaluated by blood gas analyzer in Luoyang Central Hospital Affiliated to Zhengzhou University.[Result] Imidapril significantly increased hematocrit (HCT) and total hemoglobin content (THBC) and blood Na concentration in broiler chickens exposed to low ambient temperature.No significant differences were observed in pH,Pco2,Po2,K+,Ca2+,HCO3-,HCO3std,Tco2,BE and SO2c.[Conclusion] Imidapril increases hematocrit,total hemoglobin content and blood Na+concentration in chickens exposed to low ambient temperature.

  14. A signal processing application for evaluating self-monitoring blood glucose strategies in a software agent model.

    Science.gov (United States)

    Wang, Zhanle; Paranjape, Raman

    2015-07-01

    We propose the signal processing technique of calculating a cross-correlation function and an average deviation between the continuous blood glucose and the interpolation of limited blood glucose samples to evaluate blood glucose monitoring frequency in a self-aware patient software agent model. The diabetic patient software agent model [1] is a 24-h circadian, self-aware, stochastic model of a diabetic patient's blood glucose levels in a software agent environment. The purpose of this work is to apply a signal processing technique to assist patients and physicians in understanding the extent of a patient's illness using a limited number of blood glucose samples. A second purpose of this work is to determine an appropriate blood glucose monitoring frequency in order to have a minimum number of samples taken that still provide a good understanding of the patient's blood glucose levels. For society in general, the monitoring cost of diabetes is an extremely important issue, and these costs can vary tremendously depending on monitoring approaches and monitoring frequencies. Due to the cost and discomfort associated with blood glucose monitoring, today, patients expect monitoring frequencies specific to their health profile. The proposed method quantitatively assesses various monitoring protocols (from 6 times per day to 1 time per week) in nine predefined categories of patient agents in terms of risk factors of health status and age. Simulation results show that sampling 6 times per day is excessive, and not necessary for understanding the dynamics of the continuous signal in the experiments. In addition, patient agents in certain conditions only need to sample their blood glucose 1 time per week to have a good understanding of the characteristics of their blood glucose. Finally, an evaluation scenario is developed to visualize this concept, in which appropriate monitoring frequencies are shown based on the particular conditions of patient agents. This base line can

  15. A CMOS-Based Tactile Sensor for Continuous Blood Pressure Monitoring

    CERN Document Server

    Kirstein, K -U; Salo, T; Hagleitner, C; Vancura, T; Hierlemann, A

    2011-01-01

    A monolithic integrated tactile sensor array is presented, which is used to perform non-invasive blood pressure monitoring of a patient. The advantage of this device compared to a hand cuff based approach is the capability of recording continuous blood pressure data. The capacitive, membrane-based sensor device is fabricated in an industrial CMOS-technology combined with post-CMOS micromachining. The capacitance change is detected by a S?-modulator. The modulator is operated at a sampling rate of 128kS/s and achieves a resolution of 12bit with an external decimation filter and an OSR of 128.

  16. Monitoring soil greenhouse gas emissions from managed grasslands

    Science.gov (United States)

    Díaz-Pinés, Eugenio; Lu, Haiyan; Butterbach-Bahl, Klaus; Kiese, Ralf

    2014-05-01

    Grasslands in Central Europe are of enormous social, ecological and economical importance. They are intensively managed, but the influence of different common practices (i.e. fertilization, harvesting) on the total greenhouse gas budget of grasslands is not fully understood, yet. In addition, it is unknown how these ecosystems will react due to climate change. Increasing temperatures and changing precipitation will likely have an effect on productivity of grasslands and on bio-geo-chemical processes responsible for emissions of carbon dioxide (CO2), methane (CH4) and nitrous oxide (N2O). In the frame of the TERENO Project (www.tereno.net), a long-term observatory has been implemented in the Ammer catchment, southern Germany. Acting as an in situ global change experiment, 36 big lysimeters (1 m2 section, 150 cm height) have been translocated along an altitudinal gradient, including three sites ranging from 600 to 860 meters above sea level. In addition, two treatments have been considered, corresponding to different management intensities. The overall aim of the pre-alpine TERENO observatory is improving our understanding of the consequences of climate change and management on productivity, greenhouse gas balance, soil nutritional status, nutrient leaching and hydrology of grasslands. Two of the sites are equipped with a fully automated measurement system in order to continuously and accurately monitor the soil-atmosphere greenhouse gas exchange. Thus, a stainless steel chamber (1 m2 section, 80 cm height) is controlled by a robotized system. The chamber is hanging on a metal structure which can move both vertically and horizontally, so that the chamber is able to be set onto each of the lysimeters placed on the field. Furthermore, the headspace of the chamber is connected with a gas tube to a Quantum Cascade Laser, which continuously measures CO2, CH4, N2O and H2O mixing ratios. The chamber acts as a static chamber and sets for 15 minutes onto each lysimeter

  17. 经皮测胆红素在新生儿高胆红素血症中的应用价值%The Application Value of Transcutaneous Bilirubin in Neonatal Hyperbilirubinemia

    Institute of Scientific and Technical Information of China (English)

    周琼; 张飞忠

    2014-01-01

    Objectiv:To investigate the application value of transcutaneous bilirubin (TCB) in neonatal hyperbilirubinemia. Methods:215 cases of children with hyperbilirubinemia were dynamically tested by transcutaneous bilirubin method and blood bilirubin method at 6h, 12h, 2d, 4d, 7d and other time points respectively after the admission, and the correlation of the two sets of data was compared.Results:the values measured by transcutaneous bilirubin method and blood bilirubin method at 6h, 12h, 2d, 4d, 7d and other time points showed a linear correlation relationship (P<0.05).Conclusion:TCB has many advantages such as non-invasive, repeatable testing, easy to perform, rapid detection, dynamic monitoring of neonatal bilirubin changes, etc. it deserves clinical application.%目的::探讨经皮测胆红素(TCB)在新生儿高胆红素血症中的应用价值。方法:在入院第6h、12h、2d、4d、7d等时间点采用经皮测胆红素和经血测胆红素方法对215例高胆红素血症患儿进行动态检测,比较两组数据的相关性。结果:经皮测胆红素值和经血测胆红素值在新生儿入院第6h、12h、2d、4d、7d等时间点呈直线相关性关系(P<0.05)。结论:TCB具有无创伤性、可重复多次检测、操作简单易学、检测迅速、可动态监测新生儿胆红素变化情况等优点,值得临床推广使用。

  18. Performance of the gas gain monitoring system of the CMS RPC muon detector

    CERN Document Server

    Benussi, L; Passamonti, L; Piccolo, D; Pierluigi, D; Raffone, G; Russo, A; Saviano, G; Ban, Y; Cai, J; Li, Q; Qian, S; Wang, D; Xu, Z; Zhang, F; Choi, Y; Kim, D; Choi, S; Hong, B; Kang, J W; Kang, M; Kwon, J H; Lee, K S; Park, S K; Pant, L; Singh, V B J; Kumar, A M R; Kumar, S; Chand, S; Singh, A; Bhandari, V K; Cimmino, A; Ocampo, A; Thyssen, F; Tytgat, M; Van Doninck, W; Ahmad, A; Muhamma, S; Shoaib, M; Hoorani, H; Awan, I; Ali, I; Ahmed, W; Asqhar, M I; Shahzad, H; Sayed, A; Ibrahim, A; Ali, S; Ali, R; Radi, A; Elkafrawi, T; Sharma, A; Colafranceschi, S; Abbrescia, M; Verwilligen, P; Meola, S; Cavallo, N; Braghieri, A; Montagna, P; Riccardi, C; Salvini, P; Vitulo, P; Dimitrov, A; Hadjiiska, R; Litov, L; Pavlov, B; Petkov, P; Aleksandrov, A; Genchev, V; Iaydjiev, P; Rodozov, M; Sultanov, G; Vutova, M; Stoykova, S; Ibarguen, H S; Pedraza Morales, M I; Bernardino, S Carpinteyro; Bagaturia, I

    2015-01-01

    The RPC muon detector of the CMS experiment at the LHC (CERN, Geneva, Switzerland) is equipped with a Gas Gain Monitoring (GGM) system. A report on the stability of the system during the 2011-2012 data taking run is given, as well as the observation of an effect which suggests a novel method for the monitoring of gas mixture composition.

  19. NOVAC - A global network for volcanic gas monitoring

    Science.gov (United States)

    Galle, B.

    2010-03-01

    This paper presents the global network, NOVAC (Network for Observation of Volcanic and Atmospheric Change), aiming at automatic gas emission monitoring at active volcanoes worldwide. Data from the network will primarily be used for volcanic risk assessment but also for geophysical research, studies of atmospheric change and ground validation of satellite instruments. A novel type of instrument, the Scanning miniaturized Differential Optical Absorption Spectroscopy (mini-DOAS) instrument, is applied in the network to measure volcanic gas emissions by UV absorption spectroscopy. The instrument is set up 5-10 km downwind of the volcano under study and typically 2-4 instruments are deployed at each volcano in order to cover different wind directions and facilitate measurements of plume height and plume direction. Two different versions of the instrument have been developed. Version I was designed to be a robust and simple instrument for measurement of volcanic SO2 emissions at high time-resolution with minimal power consumption. Version II was designed to allow the best possible spectroscopy, and enhanced flexibility in regard to measurement geometry at the cost of larger complexity, power consumption and price. In the paper the project is described as well as the developed software, the hardware of the two instrument versions, measurement strategies, data communication and archiving routines. As of December 2008 a total of 46 instruments have been installed at 18 volcanoes worldwide. As a typical example the installation at Tungurahua Volcano in Ecuador is described, together with some results from the first 21 months of operation at this volcano.

  20. Direct Determination of the Blood Concentration of Halogenated Anesthetic Agents by Gas Chromatography

    Directory of Open Access Journals (Sweden)

    Yamada,Teruo

    1988-08-01

    Full Text Available The direct determination by gas chromatography of blood levels of anesthetic agents has been difficult because of the water content of blood. In the present study, the method of Yokota et al. (1967 was modified by improving the packing materials of the column, the blood sample vaporizer and the flow-path during analysis. As a result, accurate and reproducible determination of halothane, enflurane and isoflurane dissolved in blood was achieved. With this system, blood in which halothane, enflurane and isoflurane had been dissolved could be analyzed without changing the column between samples. Moreover, each sample was prepared in less than 10 min, and more than 100 consecutive determinations could be made with excellent reproducibility. The coefficient of variation was less than 3.8%.

  1. Changes in electrolytes and blood gas after transfusion of irradiated MAP

    International Nuclear Information System (INIS)

    This study was undertaken to investigate the changes in recipient serum electrolytes and arterial blood gas after irradiated blood transfusion. We measured electrolytes and arterial blood gas before and after the transfusion during elective surgery in 55 patients. The mean blood loss was 1,477 g and the mean transfused blood unit of irradiated Mannitol-Adenine-Phosphate (MAP) was 5.9 units. Potassium concentration increased from 3.8 to 4.2 mEq·l-1. A total of 187 units, which had been stored for 12.3±3.5 days after donation and for 5.1±3.8 days after irradiation, was used. There was a significant correlation between the storage period from the day of irradiation and the potassium concentration of the supernatant (r=0.56, p<0.0001). This study recommends that a safe transfusion rate of irradiated blood should be determined because the high potassium concentration of irradiated blood may lead to lethal complications in case of rapid transfusion. (author)

  2. Effect of stratified inequality of blood flow on gas exchange in liquid-filled lungs.

    Science.gov (United States)

    West, J. B.; Maloney, J. E.; Castle, B. L.

    1972-01-01

    This investigation set out to answer two questions: (1) are the distal alveoli in the terminal lung units less well perfused than the proximal alveoli, i.e., is there stratification of blood flow; and (2) if so, does this enhance gas exchange in the presence of stratified inequality of ventilation. Excised dog lungs were ventilated with saline and perfused with blood. Following single inspirations of xenon 133 in saline and various periods of breath holding, the expired xenon concentration against volume was measured and it confirmed marked stratified inequality of ventilation under these conditions. By measuring the rate of depletion of xenon from alveoli during a period of blood flow, we showed that the alveoli which emptied at the end of expiration had 16% less blood flow than those exhaling earlier. However, by measuring the xenon concentration in pulmonary venous blood, we found that about 10% less tracer was transferred from the alveoli into the blood when the inspired xenon was stratified within the respiratory zone. Thus while stratification of blood flow was confirmed, it was shown to impair rather than enhance the efficiency of gas transfer.

  3. ED 03-4 HIGH-TECH IN BLOOD PRESSURE MONITORING: WHERE ARE WE GOING?

    Science.gov (United States)

    Park, Seung Woo

    2016-09-01

    Rapid progress of mobile information technologies such as wearable sensors, wireless communication, and world-wide use of smartphone cause digital health innovations. In the field of hypertension, wearable blood pressure (BP) monitoring and its wireless transfer to anywhere through smartphone, mobile smartphone apps, and cuffless blood pressure monitoring system are expected to change the way of diagnosis and management of hypertension. Home BP monitoring would be easier and wireless data transfer to health care providers would be common. Therefore, accurate diagnosis of hypertension, precise monitoring of compliance on medication, better control of BP could be achieved. In addition, BP management strategy may be changed if cuffless BP measurement becomes popular because it can monitor BP continuously and avoid sleep-disrupting by cuff inflations at night. However, all the new technologies require strict validation in terms of accuracy, cost effectiveness, and privacy to become clinically useful. It takes a lot of time and money. But speed of the technology evolvement could be too fast to be validated clinically. Consequently, new paradigm may need in the era of digital health innovation. PMID:27643038

  4. Results of vapor space monitoring of flammable gas Watch List tanks

    International Nuclear Information System (INIS)

    This report documents the measurement of headspace gas concentrations and monitoring results from the Hanford tanks that have continuous flammable gas monitoring. The systems used to monitor the tanks are Standard Hydrogen Monitoring Systems. Further characterization of the tank off-gases was done with Gas Characterization Systems and vapor grab samples. The background concentrations of all tanks are below the action level of 6250 ppm. Other information which can be derived from the measurements (such as generation rate, release rate, and ventilation rate) is also discussed

  5. Effects of insufficient sleep on blood pressure monitored by a new multibiomedical recorder.

    Science.gov (United States)

    Tochikubo, O; Ikeda, A; Miyajima, E; Ishii, M

    1996-06-01

    Blood pressure varies in relation to factors such as physical activity, body position, ambient temperature, and autonomic nervous system activity. Therefore, we have developed a portable multibiomedical (PMB) recorder that monitors five parameters: indirect blood pressure, physical activity, body position, ambient temperature, and RR interval of the electrocardiogram. In the present study, we applied the PMB recorder over a 24-hour period to study the effect of insufficient sleep on blood pressure in subjects doing extensive overtime work. The parameters listed above were measured by the PMB recorder throughout a normal workday (mean period of sleep, 8 hours) and throughout a day with insufficient sleep (mean period of sleep, 3.6 hours) in 18 male technical workers aged 23 to 48 years old. Blood pressure (mean systolic/diastolic pressure +/- SD) significantly increased the day after a sleep-insufficient night (129 +/- 8/79 +/- 6 mm Hg) compared with the day after a normal night (123 +/- 8/76 +/- 7 mm Hg, Psleep-insufficient day (2.17 +/- 0.37 versus 1.81 +/- 0.37), as was the urinary excretion of norepinephrine (Psleep-insufficient day (81 +/- ll versus 76 +/- 8 beats per minute), after the data of two subjects with abnormal levels of physical activity were excluded (Plack of sleep may increase sympathetic nervous system activity on the following day, leading to increased blood pressure. The PMB recorder was useful for precisely evaluating the relationship between blood pressure and environmental factors. PMID:8641742

  6. Non-invasive monitoring of blood pressure using the Philips Intellivue MP50 monitor cannot replace invasive blood pressure techniques in surgery patients under general anesthesia.

    Science.gov (United States)

    Meng, Xianghu; Zang, Guanghui; Fan, Longchang; Zheng, Lei; Dai, Jinzhen; Wang, Xueren; Xia, Wei; Liu, Jihong; Zhang, Chuanhan

    2013-07-01

    The Philips Intellivue MP50 monitor provides a method for non-invasive, near-continuous blood pressure (BP) monitoring and is designed to be an alternative to direct intra-arterial BP (IABP) measurement. However, no studies have specifically compared non-invasive and invasive BP measurements using the monitor. The present retrospective study observed 515 patients undergoing surgery with general anesthesia, whose invasive (intra-radial, femoral or dorsalis pedis artery) and non-invasive (oscillometric) BP (NIBP) were monitored simultaneously using the monitor. These data were analyzed using correlations, regressions and Bland-Altman plots. The patients were placed in a supine position during surgery. The correlation data for invasive BP and NIBP measurements were: for intra-radial measurements, r(2)=0.51 (bias and precision, 11.04±15.22 and 14.76±11.64 mmHg, respectively) for systolic BP (SBP) and r(2)=0.27 (6.17±11.95 and 9.77±9.25 mmHg, respectively) for diastolic BP (DBP); for intra-femoral measurements: r(2)=0.57 (14.79±14.55 and 17.15±11.68 mmHg, respectively) for SBP and r(2)=0.45 (4.12±9.70 and 7.49±7.40 mmHg, respectively) for DBP; and for intra-dorsalis pedis measurements: r(2)=0.33 (13.00±16.81 and 17.34±12.28 mmHg, respectively) for SBP and r(2)=0.30 (0.17±11.27 and 8.44±7.46 mmHg, respectively) for DBP. According to this data, the NIBP measured by the Philips Intellivue MP50 monitor showed low positive correlations and poor agreement with the IABP, as calculated by Bland-Altman analysis. Therefore, the use of oscillometric BP measured by the monitor in surgery patients under general anesthesia is not generally recommended. PMID:23935710

  7. Frequency and motives of blood glucose self-monitoring in type 1 diabetes

    DEFF Research Database (Denmark)

    Hansen, M V; Pedersen-Bjergaard, U; Heller, S R;

    2009-01-01

    AIMS: Recommendations for self-monitoring of blood glucose (SMBG) from the DCCT have not been implemented with the same rigour as recommendations for intensifying insulin therapy. We assessed the frequency of and motives for SMBG and compared SMBG behaviour with clinical, behavioural and demograp......AIMS: Recommendations for self-monitoring of blood glucose (SMBG) from the DCCT have not been implemented with the same rigour as recommendations for intensifying insulin therapy. We assessed the frequency of and motives for SMBG and compared SMBG behaviour with clinical, behavioural and...... hypoglycaemia and awareness of hypoglycaemia were independently associated with testing behaviour, whereas the presence of late diabetic complications was not. Lower HbA1c was associated with more frequent testing. CONCLUSION: Patient compliance regarding SMBG is limited. Thus, almost two thirds of the patients...

  8. Non-invasive blood glucose monitoring with Raman spectroscopy: prospects for device miniaturization

    Science.gov (United States)

    Wróbel, M. S.

    2016-01-01

    The number of patients with diabetes has reached over 350 million, and still continues to increase. The need for regular blood glucose monitoring sparks the interest in the development of modern detection technologies. One of those methods, which allows for noninvasive measurements, is Raman spectroscopy. The ability of infrared light to penetrate deep into tissues allows for obtaining measurements through the skin without its perforation. This paper presents the limitations and possibilities of non-invasive blood glucose monitoring with Raman spectroscopy. Especially focusing on the possibilities for device miniaturization. Such device incorporates a Raman spectrometer, a fiber-optical probe, and a computing device (microcontroller, smartphone, etc.) which calculates the glucose concentration using specialized algorithms. Simplification of device design, as well as turbidity correction technique and a new proposed method of synchronized detection are described.

  9. Accuracy of continuous noninvasive hemoglobin monitoring for the prediction of blood transfusions in trauma patients.

    Science.gov (United States)

    Galvagno, Samuel M; Hu, Peter; Yang, Shiming; Gao, Cheng; Hanna, David; Shackelford, Stacy; Mackenzie, Colin

    2015-12-01

    Early detection of hemorrhagic shock is required to facilitate prompt coordination of blood component therapy delivery to the bedside and to expedite performance of lifesaving interventions. Standard physical findings and vital signs are difficult to measure during the acute resuscitation stage, and these measures are often inaccurate until patients deteriorate to a state of decompensated shock. The aim of this study is to examine a severely injured trauma patient population to determine whether a noninvasive SpHb monitor can predict the need for urgent blood transfusion (universal donor or additional urgent blood transfusion) during the first 12 h of trauma patient resuscitation. We hypothesize that trends in continuous SpHb, combined with easily derived patient-specific factors, can identify the immediate need for transfusion in trauma patients. Subjects were enrolled if directly admitted to the trauma center, >17 years of age, and with a shock index (heart rate/systolic blood pressure) >0.62. Upon admission, a Masimo Radical-7 co-oximeter sensor (Masimo Corporation, Irvine, CA) was applied, providing measurement of continuous non-invasive hemoglobin (SpHb) levels. Blood was drawn and hemoglobin concentration analyzed and conventional pulse oximetry photopletysmograph signals were continuously recorded. Demographic information and both prehospital and admission vital signs were collected. The primary outcome was transfusion of at least one unit of packed red blood cells within 24 h of admission. Eight regression models (C1-C8) were evaluated for the prediction of blood use by comparing area under receiver operating curve (AUROC) at different time intervals after admission. 711 subjects had continuous vital signs waveforms available, to include heart rate (HR), SpHb and SpO2 trends. When SpHb was monitored for 15 min, SpHb did not increase AUROC for prediction of transfusion. The highest ROC was recorded for model C8 (age, sex, prehospital shock index, admission

  10. Generalized model of synthesis quality monitoring system of extraction, processing and transportation of gas

    Directory of Open Access Journals (Sweden)

    Леонид Иванович Нефедов

    2014-08-01

    Full Text Available The extraction and processing and gas transportation have been analyzed in the article. Decomposition of the process of monitoring is introduced based on that; allowing the hierarchical structure of the monitoring system is request. Scientific novelty consists in the fact that developed generalized synthesis quality model monitoring system to extract, processing and gas transportation, which allows solving the problem of synthesis of a unified system and criteria positions.

  11. A Microwave Metamaterial Inspired Sensor for Non-Invasive Blood Glucose Monitoring

    Directory of Open Access Journals (Sweden)

    J. Vrba

    2015-12-01

    Full Text Available In this paper, a microwave sensor based on an artificial transmission line is proposed for non-invasive blood glucose monitoring. A corresponding numerical model of the sensor implemented in microstrip technology is created in the commercial full-wave numerical simulation tool COMSOL Multiphysics and virtually tested by means of numerical simulations. Blood-glucose solution models with various blood glucose concentrations are used as a model of a biological tissue under test. Furthermore, a possible methodology for performing non-invasive tests is proposed. Sensitivity of the sensor developed here is compared to a sensor based on a section of a conventional microstrip transmission line of the same length and width.

  12. Doppler standard deviation imaging for clinical monitoring of in vivo human skin blood flow

    Energy Technology Data Exchange (ETDEWEB)

    Zhao, Yonghua; Chen, Zhongping; Saxer, Christopher; Shen, Qimin; Xiang, Shaohua; Boer, Johannes F. de; Nelson, J. Stuart

    2000-09-15

    We used a novel phase-resolved optical Doppler tomographic (ODT) technique with very high flow-velocity sensitivity (10 {mu}m/s) and high spatial resolution (10 {mu}m) to image blood flow in port-wine stain (PWS) birthmarks in human skin. In addition to the regular ODT velocity and structural images, we use the variance of blood flow velocity to map the PWS vessels. Our device combines ODT and therapeutic systems such that PWS blood flow can be monitored in situ before and after laser treatment. To the authors' knowledge this is the first clinical application of ODT to provide a fast semiquantitative evaluation of the efficacy of PWS laser therapy in situ and in real time. (c) 2000 Optical Society of America.

  13. A wearable diffuse reflectance sensor for continuous monitoring of cutaneous blood content

    Energy Technology Data Exchange (ETDEWEB)

    Zakharov, P; Talary, M S; Caduff, A [Solianis Monitoring AG, Leutschenbachstrasse 46, CH-8050 Zuerich (Switzerland)], E-mail: andreas.caduff@solianis.com

    2009-09-07

    An optical diffuse reflectance sensor for characterization of cutaneous blood content and optimized for continuous monitoring has been developed as part of a non-invasive multisensor system for glucose monitoring. A Monte Carlo simulation of the light propagation in the multilayered skin model has been performed in order to estimate the optimal geometrical separation of the light source and detector for skin and underlying tissue. We have observed that the pathlength within the upper vascular plexus of the skin which defines the sensor sensitivity initially grows with increasing source-detector distance (SDD) before reaching a maximum at 3.5 mm and starts to decay with further increase. At the same time, for distances above 2.4 mm, the sensor becomes sensitive to muscle blood content, which decreases the specificity to skin perfusion monitoring. Thus, the SDDs in the range from 1.5 mm to 2.4 mm satisfy the requirements of sensor sensitivity and specificity. The hardware implementation of the system has been realized and tested in laboratory experiments with a venous occlusion procedure and in an outpatient clinical study in 16 patients with type 1 diabetes mellitus. For both testing procedures, the optical sensor demonstrated high sensitivity to perfusion change provoking events. The general build-up of cutaneous blood under the sensor has been observed which can be associated with pressure-induced vasodilation as a response to the sensor application.

  14. Transcutaneous vs. intraoperative quantitative ultrasound for staging bovine hepatic steatosis.

    NARCIS (Netherlands)

    Weijers, G.; Starke, A.; Thijssen, J.M.; Haudum, A.; Wohlsein, P.; Rehage, J.; Korte, C.L. de

    2012-01-01

    The aim of this study was to test the hypothesis that quantitative analysis of transcutaneous (Transc) ultrasound (US) images can predict the liver fat content with similar accuracy and precision as using intraoperative (Intraop) US. The second goal was to investigate if a tissue mimicking phantom (

  15. Blood flow changes after unilateral carotid artery ligation monitored by optical coherence tomography

    Science.gov (United States)

    Ma, Yushu; Liang, Chengbo; Suo, Yanyan; Zhao, Yuqian; Wang, Yi; Xu, Tao; Wang, Ruikang; Ma, Zhenhe

    2016-03-01

    Unilateral carotid artery ligation which could induce adaptive improvement is a classic model that has been widely used to study pathology of ischemic disease. In those studies, blood flow is an important parameter to characterize the ischemia. Optical coherence tomography (OCT) is a powerful imaging modality which can provide depth resolved images in biological tissue with high spatial and temporal resolution. SPF rats was anesthetized with isoflurane and divided into two groups. In first group, bilateral carotid artery was surgically exposed, and then left carotid artery was ligated. Blood flow changes of the contralateral carotid artery was monitored using high speed spectral domain optical coherence tomography, including the absolute flow velocity and the flow volume. In the other group, skull window was opened at the ipsilateral cerebral cortex of ligation and blood supply of small artery was measured before and after the ligation. The measured results demonstrate the blood supply compensation process after unilateral carotid artery ligation. With the superiority of high resolution, OCT is an effective technology in monitoring results of carotid artery after ligation.

  16. Implantable sensor for intraoperative and postoperative monitoring of blood flow: a preliminary report.

    Science.gov (United States)

    Rabinovitz, R S; Hartley, C J; Michael, L H; Entman, M L; Hawkins, H K; Sekela, M E; Noon, G P

    1990-08-01

    We have developed an implantable 20 MHz Doppler flow sensor, to monitor blood flow in patients after operation. During surgery the sensor is wrapped around a blood vessel and secured in place with a releasable tie. No tissue puncturing techniques are required. The lead wires, together with a release cable, are exteriorized through the chest wall. Several days after operation, the tie is externally released, and the sensor is pulled out. Twenty-seven sensors were implanted in 24 dogs for up to 16 days. All were extracted successfully with minimal visible behavioral reaction in the awake dogs and caused no thrombosis or vascular damage. The Doppler velocity signals had excellent linear correlations (r = 0.99) with data from electromagnetic flow sensors and timed blood collections. In addition, the sensors were applied to coronary artery bypass grafts in 31 patients for up to 2 days after operation. High-quality signals were obtained, and intraoperatively recorded signals agreed well with simultaneous electromagnetic flow tracings. All sensors were extracted with minimal discomfort to the awake patients and with no complications. Thus, the implantable flow sensor is a safe, reliable, accurate, and simple method for intraoperative and postoperative monitoring of blood flow in vessels 2.5 to 5 mm in diameter. PMID:2199684

  17. Portable blood gas and electrolyte analyzer evaluated in a multiinstitutional study

    NARCIS (Netherlands)

    J. Lindemans (Jan); P. Hoefkens (Peter); A.L. van Kessel; M. Bonnay; W.R. Kulpmann; J.D.E. van Suijlen

    1999-01-01

    textabstractA recently introduced blood gas/electrolyte analyzer (SenDx 100((R)), renamed ABL70) intended for point-of-care, near-patient, or stat laboratory use was evaluated simultaneously in four different institutions and compared with three different laboratory ben

  18. Inert tracer gas washout from mixed venous blood: the sloping alveolar plateau

    NARCIS (Netherlands)

    Vries, W.R. de; Luijendijk, S.C.M.; Zwart, Aart

    1982-01-01

    The aim of this model study was to investigate the mechanisms underlying the sloping alveolar plateau for inert tracer gases supplied to the lung by mixed venous blood. Transpulmonary gas exchange was simulated in an asymmetric lung model for conditions at rest and in exercise. For highly soluble ga

  19. New Criteria for Assessing the Accuracy of Blood Glucose Monitors meeting, October 28, 2011.

    Science.gov (United States)

    Walsh, John; Roberts, Ruth; Vigersky, Robert A; Schwartz, Frank

    2012-03-01

    Glucose meters (GMs) are routinely used for self-monitoring of blood glucose by patients and for point-of-care glucose monitoring by health care providers in outpatient and inpatient settings. Although widely assumed to be accurate, numerous reports of inaccuracies with resulting morbidity and mortality have been noted. Insulin dosing errors based on inaccurate GMs are most critical. On October 28, 2011, the Diabetes Technology Society invited 45 diabetes technology clinicians who were attending the 2011 Diabetes Technology Meeting to participate in a closed-door meeting entitled New Criteria for Assessing the Accuracy of Blood Glucose Monitors. This report reflects the opinions of most of the attendees of that meeting. The Food and Drug Administration (FDA), the public, and several medical societies are currently in dialogue to establish a new standard for GM accuracy. This update to the FDA standard is driven by improved meter accuracy, technological advances (pumps, bolus calculators, continuous glucose monitors, and insulin pens), reports of hospital and outpatient deaths, consumer complaints about inaccuracy, and research studies showing that several approved GMs failed to meet FDA or International Organization for Standardization standards in postapproval testing. These circumstances mandate a set of new GM standards that appropriately match the GMs' analytical accuracy to the clinical accuracy required for their intended use, as well as ensuring their ongoing accuracy following approval. The attendees of the New Criteria for Assessing the Accuracy of Blood Glucose Monitors meeting proposed a graduated standard and other methods to improve GM performance, which are discussed in this meeting report.

  20. Measurement of tissue optical properties with optical coherence tomography: Implication for noninvasive blood glucose concentration monitoring

    Science.gov (United States)

    Larin, Kirill V.

    Approximately 14 million people in the USA and more than 140 million people worldwide suffer from diabetes mellitus. The current glucose sensing technique involves a finger puncture several times a day to obtain a droplet of blood for analysis. There have been enormous efforts by many scientific groups and companies to quantify glucose concentration noninvasively using different optical techniques. However, these techniques face limitations associated with low sensitivity, accuracy, and insufficient specificity of glucose concentrations over a physiological range. Optical coherence tomography (OCT), a new technology, is being applied for noninvasive imaging in tissues with high resolution. OCT utilizes sensitive detection of photons coherently scattered from tissue. The high resolution of this technique allows for exceptionally accurate measurement of tissue scattering from a specific layer of skin compared with other optical techniques and, therefore, may provide noninvasive and continuous monitoring of blood glucose concentration with high accuracy. In this dissertation work I experimentally and theoretically investigate feasibility of noninvasive, real-time, sensitive, and specific monitoring of blood glucose concentration using an OCT-based biosensor. The studies were performed in scattering media with stable optical properties (aqueous suspensions of polystyrene microspheres and milk), animals (New Zealand white rabbits and Yucatan micropigs), and normal subjects (during oral glucose tolerance tests). The results of these studies demonstrated: (1) capability of the OCT technique to detect changes in scattering coefficient with the accuracy of about 1.5%; (2) a sharp and linear decrease of the OCT signal slope in the dermis with the increase of blood glucose concentration; (3) the change in the OCT signal slope measured during bolus glucose injection experiments (characterized by a sharp increase of blood glucose concentration) is higher than that measured in

  1. Improving the gas gain monitoring system in multiwire proportional chambers for MUON detector of LHCb experiment.

    CERN Document Server

    Ruvinskaia, Ekaterina

    2016-01-01

    The gas gain monitoring system in multi-wire proportional chambers for MUON detector of LHCb has been constructed and commissioned. It includes an online- monitoring, tools for analysis the archived data and an alarm system on the quality of the gas mixture. Finally, it will be implemented in the main ECS of LHCb for MUON detector and as a part of safety system of LHCb as a permanent online monitor of the quality of the gas mixture in MWPCs. The main advantage of this setup is a monitoring of Gas Gain (GG) in MWPCs with radioactive sources independently from the presence of beam at LHC. It also provides an option for prompt reaction in case of a problem with the gas.

  2. Monitoring Gas Void Fraction In Two-Phase Flow With Acoustic Emission

    OpenAIRE

    Addali, Abdulmajid

    2010-01-01

    The two-phase gas/liquid flow phenomenon can be encountered over a range of gas and liquid flow rates in the chemical engineering industry, particularly in oil and gas production transportation pipelines. Monitoring and measurement of their characteristics, such as the gas void fraction, are necessary to minimise the disruption of downstream process facilities. Thus, over the last decade, the investigation, development and use of multiphase flow metering system have been a major focus for the...

  3. Results of ambulatory arterial blood pressure monitoring in children with obesity

    Directory of Open Access Journals (Sweden)

    Faruk Öktem

    2010-12-01

    Full Text Available Objectives: The relationship between obesity and essential hypertension is well known. In this study, we aimed to evaluate ambulatory arterial blood pressure monitoring of obese and non-obese children who had similar demographic characteristics.Materials and methods: Seventy one children and adolescents (n=39 obesity, n=32 controls were studied. Blood pressure of the children were measured by 24 hour ambulatory blood pressure monitoring device.Results: Obese children had significantly higher mean blood pressure values (systolic 121.9±11.7 mmHg, diastolic 70.2±5.3 mmHg than control subjects (systolic 109.3±6.7 mmHg, diastolic 65.1±4.6 mmHg, p0.05. Blood pressure load was found to be increased in obese children compared to the controls (%13.6±12.9 and %2.6±3.4, respectively; p<0.05. Serum total cholesterol and LDL-cholesterol levels of obese children (181.1±33.4 and 131.1±23.1mg/dl were significantly higher than those of the controls (134.3±11.1 and 103.3±14.2 mg/dl, p<0.05.Conclusions: Obesity in children and adolescents should not be regarded as variations of normality, but as abnormality with an extremely high risk for the development of hypertension and hyperlipidemia in adulthood.

  4. Systemic Metabolomic Changes in Blood Samples of Lung Cancer Patients Identified by Gas Chromatography Time-of-Flight Mass Spectrometry

    Directory of Open Access Journals (Sweden)

    Suzanne Miyamoto

    2015-04-01

    Full Text Available Lung cancer is a leading cause of cancer deaths worldwide. Metabolic alterations in tumor cells coupled with systemic indicators of the host response to tumor development have the potential to yield blood profiles with clinical utility for diagnosis and monitoring of treatment. We report results from two separate studies using gas chromatography time-of-flight mass spectrometry (GC-TOF MS to profile metabolites in human blood samples that significantly differ from non-small cell lung cancer (NSCLC adenocarcinoma and other lung cancer cases. Metabolomic analysis of blood samples from the two studies yielded a total of 437 metabolites, of which 148 were identified as known compounds and 289 identified as unknown compounds. Differential analysis identified 15 known metabolites in one study and 18 in a second study that were statistically different (p-values <0.05. Levels of maltose, palmitic acid, glycerol, ethanolamine, glutamic acid, and lactic acid were increased in cancer samples while amino acids tryptophan, lysine and histidine decreased. Many of the metabolites were found to be significantly different in both studies, suggesting that metabolomics appears to be robust enough to find systemic changes from lung cancer, thus showing the potential of this type of analysis for lung cancer detection.

  5. Real-time electrical impedimetric monitoring of blood coagulation process under temperature and hematocrit variations conducted in a microfluidic chip.

    Directory of Open Access Journals (Sweden)

    Kin Fong Lei

    Full Text Available Blood coagulation is an extremely complicated and dynamic physiological process. Monitoring of blood coagulation is essential to predict the risk of hemorrhage and thrombosis during cardiac surgical procedures. In this study, a high throughput microfluidic chip has been developed for the investigation of the blood coagulation process under temperature and hematocrit variations. Electrical impedance of the whole blood was continuously recorded by on-chip electrodes in contact with the blood sample during coagulation. Analysis of the impedance change of the blood was conducted to investigate the characteristics of blood coagulation process and the starting time of blood coagulation was defined. The study of blood coagulation time under temperature and hematocrit variations was shown a good agreement with results in the previous clinical reports. The electrical impedance measurement for the definition of blood coagulation process provides a fast and easy measurement technique. The microfluidic chip was shown to be a sensitive and promising device for monitoring blood coagulation process even in a variety of conditions. It is found valuable for the development of point-of-care coagulation testing devices that utilizes whole blood sample in microliter quantity.

  6. Use of paravascular admittance waveforms to monitor relative change in arterial blood pressure

    Science.gov (United States)

    Zielinski, Todd M.; Hettrick, Doug; Cho, Yong

    2010-04-01

    Non-invasive methods to monitor ambulatory blood pressure often have limitations that can affect measurement accuracy and patient adherence [1]. Minimally invasive measurement of a relative blood pressure surrogate with an implantable device may provide a useful chronic diagnostic and monitoring tool. We assessed a technique that uses electrocardiogram and paravascular admittance waveform morphology analysis to one, measure a time duration (vascular tone index, VTI in milliseconds) change from the electrocardiogram R-wave to admittance waveform peak and two, measure the admittance waveform minimum, maximum and magnitude as indicators of change in arterial compliance/distensibility or pulse pressure secondary to change in afterload. Methods: Five anesthetized domestic pigs (32 ± 4.2 kg) were used to study the effects of phenylephrine (1-5 ug/kg/min) on femoral artery pressure and admittance waveform morphology measured with a quadrapolar electrode array catheter placed next to the femoral artery to assess the relative change in arterial compliance due to change in peripheral vascular tone. Results: Statistical difference was observed (p blood pressure may be suitable for implantable devices to detect progression of cardiovascular disease such as hypertension.

  7. Monitoring of glucose, salt and pure water in human whole blood: An in vitro study.

    Science.gov (United States)

    Imran, Muhammad; Ullah, Hafeez; Akhtar, Munir; Sial, Muhammad Aslam; Ahmed, Ejaz; Durr-E-Sabeeh; Ahmad, Mukhtar; Hussain, Fayyaz

    2016-07-01

    Designing and implementation of non-invasive methods for glucose monitoring in blood is main focus of biomedical scientists to provide a relief from skin puncturing of diabete patient. The objective of this research work is to investigate the shape deformations and the aggregation of red blood cells (RBCs) in the human blood after addition of three different analytes i) (0mM-400mM: Range) of glucose (C(6)H(12)O(6)), ii) (0mM-400mM: range) of pure salt (NaCl) and iii) (0mM- 350mM: range) of pure water (H(2)O). We have observed that the changes in the shape of individual cells from biconcave discs to spherical shapes and eventually the lysis of the cells at optimum concentration of glucose, salts and pure water. This demonstration also provides a base line to facilitate diabetes during partial diagnosis and monitoring of the glucose levels qualitatively both in research laboratories and clinical environment. PMID:27393437

  8. Patterns of self-monitoring of blood glucose in insulin-treated diabetes: analysis of a Scottish population over time.

    Science.gov (United States)

    Cameron, D; Harris, F M; Evans, J M M

    2016-07-01

    Analysis of a diabetes clinical information system in Tayside, Scotland, shows that a significant proportion of insulin-treated patients with diabetes are not self-monitoring blood glucose according to current clinical guidance and recommendations, with some not self-monitoring their blood glucose at all. Although there has been an increase in the number of reagent strips dispensed over the past decade, this increase is mainly accounted for by increased testing frequency among people with diabetes already testing. PMID:26990502

  9. System accuracy of blood glucose monitoring systems: impact of use by patients and ambient conditions.

    Science.gov (United States)

    Schmid, Christina; Haug, Cornelia; Heinemann, Lutz; Freckmann, Guido

    2013-10-01

    For self-monitoring of blood glucose by people with diabetes, the reliability of the measured blood glucose values is a prerequisite in order to ensure correct therapeutic decisions. Requirements for system accuracy are defined by the International Organization for Standardization (ISO) in the standard EN ISO 15197:2003. However, even a system with high analytical quality is not a guarantee for accurate and reliable measurement results. Under routine life conditions, blood glucose measurement results are affected by several factors. First, the act of performing measurements as well as the handling of the system may entail numerous possible error sources, such as traces of glucose-containing products on the fingertips, the use of deteriorated test strips, or the incorrect storage of test strips. Second, ambient and sampling conditions such as high altitude, partial pressure of oxygen, ambient temperature, and the use of alternate test sites can have an influence on measurement results. Therefore, the user-friendliness of a system and the quality of the manufacturer's labeling to reduce the risk of handling errors are also important aspects in ensuring reliable and accurate measurement results. In addition, the analytical performance of systems should be less prone to user errors and ambient conditions. Finally, people with diabetes must be aware of the information and instructions in the manufacturer's labeling and must be able to measure and interpret blood glucose results correctly. PMID:23883407

  10. The determination of phenazone in blood plasma for obtained sistem suitable test of monitoring drug level

    Directory of Open Access Journals (Sweden)

    Mochamad Lazuardi

    2007-09-01

    Full Text Available The determining of Phenazone to human blood plasma from healthy man after separated by solid phase extraction (SPE and spectroscopic measurements has been investigated. The objective of that research was to obtain system suitable test for determine the Phenazone level in biological fluids (human blood plasma, for new performed dosage regimented in clinical dentistry. The method can be divided into the following four steps. 1. Centrifugation the blood sample, 2. Extraction from blood plasma and, 3. Separation by SPE with manual pressured, 4. Elution to SPE followed by the measurement on a spectrophotometer in the ultra violet region. The critical value of  │t │at the 5% confidence level indicates that there is no systematic error in the linearity proposed method. Recoveries for this research were obtained at ranging 93.460 to 95.598%. The coefficient variation precision of this procedure was clearly good at smallest than 2%. The analytical procedure can be carried out in one working operation as a monitored therapeutic activity.

  11. Home blood pressure monitor use in patients with chronic kidney disease.

    Science.gov (United States)

    Ye, Xiaojing; Shafi, Salman; Negrete, Andrea; Davis, William N; Sarac, Erdal; Negrete, Alejandro M; Negrete, Hilmer

    2016-10-01

    Home blood pressure monitoring (HBPM) is recommended in patients with chronic kidney disease (CKD) and hypertension (HTN). However, little is known about the use and pattern of HBPM in CKD patients. A cross-sectional study was conducted in an out-patient nephrology clinic. A total of 285 patients participated in the study. Of all patients, 66% reported using HBPM. Self-reported compliance with BP medications (93.9% vs. 85.1%, p 0.03), exercise >3 days/week (45.9 vs. 26.3%, p helpful in managing hypertension (85.4%). Most common reason for not using HBPM was lack of advice by a physician (43.4%). HBPM use is common in patients with CKD and HTN. HBPM users are more likely to follow life-style and dietary modifications for blood pressure control. PMID:27104796

  12. An intelligent system for continuous blood pressure monitoring on remote multi-patients in real time

    CERN Document Server

    Marani, Roberto

    2012-01-01

    In this paper we present an electronic system to perform a non-invasive measurement of the blood pressure based on the oscillometric method, which does not suffer from the limitations of the well-known auscultatory one. Moreover the proposed system is able to evaluate both the systolic and diastolic blood pressure values and makes use of a microcontroller and a Sallen-Key active filter. With reference to other similar devices, a great improvement of our measurement system is achieved since it performs the transmission of the systolic and diastolic pressure values to a remote computer. This aspect is very important when the simultaneous monitoring of multi-patients is required. The proposed system, prototyped and tested at the Electron Devices Laboratory (Electrical and Information Engineering Department) of Polytechnic University of Bari, Italy, is characterized by originality, by plainness of use and by a very high level of automation (so called intelligent system).

  13. Impact of antihypertensive combination and monotreatments on blood pressure variability: assessment by old and new indices. Data from a large ambulatory blood pressure monitoring database.

    LENUS (Irish Health Repository)

    Parati, Gianfranco

    2014-06-01

    High 24-h ambulatory blood pressure (ABP) variability is associated with poor cardiovascular outcomes. We analysed a large ABP monitoring database containing data from hypertensive patients treated with telmisartan\\/amlodipine combination or various monotherapies with the aim of quantifying the 24-h distribution of blood pressure (BP) reduction by treatment through the smoothness index and of developing and testing a new treatment-on-variability index (TOVI) to quantify the effects of treatment on both mean BP and BP variability.

  14. Combustion/Emission Species Monitoring Ground and Flight Aeronautical Research Using a Gas Microsensor Array Project

    Data.gov (United States)

    National Aeronautics and Space Administration — The goal of this program is to develop a miniaturized and in-situ operated gas microsensor array for the real time monitoring of chemical composition of turbine...

  15. Application of optical coherence tomography for assessment of transcutaneous vaccine delivery

    Science.gov (United States)

    Kamali, T.; Rattanapak, T.; Hook, S.; Meglinski, I.

    2012-03-01

    Immunization is one of the most efficient and cost-effective means for the prevention of diseases, but most vaccines have to be administered invasively. A novel strategy of inducing an immune response is topical application of vaccines to intact skin. Apart from being a non-invasive route of drug delivery, skin delivery also offers an advantageous mode of immunization due to the ability of skin immune cells to present antigens to the immune system. Topical vaccine penetration through the outermost layers of skin is based on the percutaneous diffusion of lipid-based nano-particles. In the current study we investigate the applicability of Optical Coherence Tomography for monitoring transcutaneous delivery of a peptide vaccine into the skin in vivo.

  16. [A new method for the transcutaneous measurement of deep body temperature during anaesthesia and intensive care (author's transl)].

    Science.gov (United States)

    Jost, U; Hanf, K; Köhler, C O; Just, O H

    1978-04-01

    A new method for monitoring deep body temperature is described. It is based on the establishment, by means of electronic appliances, of a zone without heatflow from the deep tissues. The method is simple and the results compare favourably with those obtained by other procedures for measuring core temperature. The uses of this transcutaneous mehtod are discussed and its advantages and reliability in the operating theatre and intensive care unit are emphasized. It becomes less reliable if it is employed during and after extracorporeal circulation in hypothermia on account of the temperature gradient.

  17. Integrated Self-Monitoring of Blood Glucose System: Handling Step Analysis

    OpenAIRE

    Freckmann, Guido; Schmid, Christina; Ruhland, Katharina; Baumstark, Annette; Haug, Cornelia

    2012-01-01

    Self-monitoring of blood glucose (SMBG) implicates a number of handling steps with the meter and the lancing device. Numerous user errors can occur during SMBG, and each step adds to the complexity of use. This report compares the required steps to perform SMBG of one fully integrated (the second generation of the Accu-Chek® Mobile), three partly integrated (Accu-Chek Compact Plus, Ascensia® Breeze®2, and Accu-Chek Aviva), and six conventional (Bayer Contour®, Bayer Contour USB, BGStar™, Free...

  18. Frequency and motives of blood glucose self-monitoring in type 1 diabetes

    DEFF Research Database (Denmark)

    Hansen, M.V.; Pedersen-Bjergaard, U.; Heller, S.R.;

    2009-01-01

    AIMS: Recommendations for self-monitoring of blood glucose (SMBG) from the DCCT have not been implemented with the same rigour as recommendations for intensifying insulin therapy. We assessed the frequency of and motives for SMBG and compared SMBG behaviour with clinical, behavioural and...... hypoglycaemia and awareness of hypoglycaemia were independently associated with testing behaviour, whereas the presence of late diabetic complications was not. Lower HbA1c was associated with more frequent testing. CONCLUSION: Patient compliance regarding SMBG is limited. Thus, almost two thirds of the patients...

  19. Transcutaneous Noninvasive Device for the Responsive Delivery of Melatonin in Microgravity. Project

    Data.gov (United States)

    National Aeronautics and Space Administration — Our goal is develop a smart, transcutaneous device for individualized circadian (sleep) therapy by responsive release of melatonin, in microgravity. Additionally,...

  20. Assessment of transcutaneous vaccine delivery by optical coherence tomography Assessment of transcutaneous vaccine delivery by OCT

    Science.gov (United States)

    Kamali, T.; Doronin, A.; Rattanapak, T.; Hook, S.; Meglinski, I.

    2012-08-01

    Immunization is one of the most efficient and cost-effective means for the prevention of diseases. The latest trend for inducing protective immunity is topical application of vaccines to intact skin rather than invasive administration via injection. Apart from being a non-invasive route of drug delivery, skin itself also offers advantages through the presence of cells of the immune system in both the dermis and epidermis. However, vaccine penetration through the outermost layers of skin is limited by the barrier provided by the Stratum corneum. In the current study utilizing conventional Optical Coherence Tomography (OCT) we investigate the transcutaneous delivery of a nano- particulate peptide vaccine into mouse skin in vivo. We demonstrate that a front of molecular diffusion within the skin can be clearly observed by using cross-correlations of successive 2D OCT images. Thus, OCT provides a unique tool for quantitative assessment of dynamics of diffusion of drugs, target compounds, analytes, cosmetics and various chemical agents in biological tissues in vivo.

  1. Long-term autonomous volcanic gas monitoring with Multi-GAS at Mount St. Helens, Washington, and Augustine Volcano, Alaska

    Science.gov (United States)

    Kelly, P. J.; Ketner, D. M.; Kern, C.; Lahusen, R. G.; Lockett, C.; Parker, T.; Paskievitch, J.; Pauk, B.; Rinehart, A.; Werner, C. A.

    2015-12-01

    In recent years, the USGS Volcano Hazards Program has worked to implement continuous real-time in situ volcanic gas monitoring at volcanoes in the Cascade Range and Alaska. The main goal of this ongoing effort is to better link the compositions of volcanic gases to other real-time monitoring data, such as seismicity and deformation, in order to improve baseline monitoring and early detection of volcanic unrest. Due to the remote and difficult-to-access nature of volcanic-gas monitoring sites in the Cascades and Alaska, we developed Multi-GAS instruments that can operate unattended for long periods of time with minimal direct maintenance from field personnel. Our Multi-GAS stations measure H2O, CO2, SO2, and H2S gas concentrations, are comprised entirely of commercial off-the-shelf components, and are powered by small solar energy systems. One notable feature of our Multi-GAS stations is that they include a unique capability to perform automated CO2, SO2, and H2S sensor verifications using portable gas standards while deployed in the field, thereby allowing for rigorous tracking of sensor performances. In addition, we have developed novel onboard data-processing routines that allow diagnostic and monitoring data - including gas ratios (e.g. CO2/SO2) - to be streamed in real time to internal observatory and public web pages without user input. Here we present over one year of continuous data from a permanent Multi-GAS station installed in August 2014 in the crater of Mount St. Helens, Washington, and several months of data from a station installed near the summit of Augustine Volcano, Alaska in June 2015. Data from the Mount St. Helens Multi-GAS station has been streaming to a public USGS site since early 2015, a first for a permanent Multi-GAS site. Neither station has detected significant changes in gas concentrations or compositions since they were installed, consistent with low levels of seismicity and deformation.

  2. The Effect of Blood Glucose Self-Monitoring Among Inmates With Diabetes.

    Science.gov (United States)

    Hunter Buskey, Robin N; Mathieson, Kathleen; Leafman, Joan S; Feinglos, Mark N

    2015-10-01

    The increasing prevalence and risk of complications from diabetes necessitate patient participation and attentiveness to select appropriate foods, perform regular physical activity, and be active in diabetes management and self-maintenance. Diabetes is often largely asymptomatic; consequently, early diagnosis and treatment are necessary. Inmates are a unique population challenged by the increased prevalence of chronic conditions including diabetes. Diabetes standards for inmates contain diagnostic and treatment management guidelines that incorporate personal glucose monitoring for insulin users. In December 2009, the Federal Bureau of Prisons initiated a program to distribute glucose meters to insulin-dependent inmates to facilitate self-monitoring blood glucose. The purpose of this study was to evaluate the effect of these glucose meters on hemoglobin A1c levels. PMID:26276137

  3. Cerebral venous blood oxygenation monitoring during hyperventilation in healthy volunteers with a novel optoacoustic system

    Science.gov (United States)

    Petrov, Andrey; Prough, Donald S.; Petrov, Irene Y.; Petrov, Yuriy; Deyo, Donald J.; Henkel, Sheryl N.; Seeton, Roger; Esenaliev, Rinat O.

    2013-03-01

    Monitoring of cerebral venous oxygenation is useful to facilitate management of patients with severe or moderate traumatic brain injury (TBI). Prompt recognition of low cerebral venous oxygenation is a key to avoiding secondary brain injury associated with brain hypoxia. In specialized clinical research centers, jugular venous bulb catheters have been used for cerebral venous oxygenation monitoring and have demonstrated that oxygen saturation capnography. Good temporal correlation between decreases in optoacoustically measured SSS oxygenation and decreases in EtCO2 was obtained. Decreases in EtCO2 from normal values (35-45 mmHg) to 20-25 mmHg resulted in SSS oxygenation decreases by 3-10%. Intersubject variability of the responses may relate to nonspecific brain activation associated with voluntary hyperventilation. The obtained data demonstrate the capability of the optoacoustic system to detect in real time minor changes in the SSS blood oxygenation.

  4. Monitoring gas emissions can help forecast volcanic eruptions

    Science.gov (United States)

    Kern, Christoph; Maarten de Moor,; Bo Galle,

    2015-01-01

    As magma ascends in active volcanoes, dissolved volatiles partition from melt into a gas phase, rise, and are released into the atmosphere from volcanic vents. The major components of high-temperature volcanic gas are typically water vapor, carbon dioxide, and sulfur dioxide. 

  5. Graphical arterial blood gas visualization tool supports rapid and accurate data interpretation.

    Science.gov (United States)

    Doig, Alexa K; Albert, Robert W; Syroid, Noah D; Moon, Shaun; Agutter, Jim A

    2011-04-01

    A visualization tool that integrates numeric information from an arterial blood gas report with novel graphics was designed for the purpose of promoting rapid and accurate interpretation of acid-base data. A study compared data interpretation performance when arterial blood gas results were presented in a traditional numerical list versus the graphical visualization tool. Critical-care nurses (n = 15) and nursing students (n = 15) were significantly more accurate identifying acid-base states and assessing trends in acid-base data when using the graphical visualization tool. Critical-care nurses and nursing students using traditional numerical data had an average accuracy of 69% and 74%, respectively. Using the visualization tool, average accuracy improved to 83% for critical-care nurses and 93% for nursing students. Analysis of response times demonstrated that the visualization tool might help nurses overcome the "speed/accuracy trade-off" during high-stress situations when rapid decisions must be rendered. Perceived mental workload was significantly reduced for nursing students when they used the graphical visualization tool. In this study, the effects of implementing the graphical visualization were greater for nursing students than for critical-care nurses, which may indicate that the experienced nurses needed more training and use of the new technology prior to testing to show similar gains. Results of the objective and subjective evaluations support the integration of this graphical visualization tool into clinical environments that require accurate and timely interpretation of arterial blood gas data.

  6. Performance Evaluation and Labeling Comprehension of a New Blood Glucose Monitoring System with Integrated Information Management

    Science.gov (United States)

    List, Susan M; Starks, Nykole; Baum, John; Greene, Carmine; Pardo, Scott; Parkes, Joan L; Schachner, Holly C; Cuddihy, Robert

    2011-01-01

    Background This study evaluated performance and product labeling of CONTOUR® USB, a new blood glucose monitoring system (BGMS) with integrated diabetes management software and a universal serial bus (USB) port, in the hands of untrained lay users and health care professionals (HCPs). Method Subjects and HCPs tested subject's finger stick capillary blood in parallel using CONTOUR USB meters; deep finger stick blood was tested on a Yellow Springs Instruments (YSI) glucose analyzer for reference. Duplicate results by both subjects and HCPs were obtained to assess system precision. System accuracy was assessed according to International Organization for Standardization (ISO) 15197:2003 guidelines [within ±15 mg/dl of mean YSI results (samples <75 mg/dl) and ±20% (samples ≥75 mg/dl)]. Clinical accuracy was determined by Parkes error grid analysis. Subject labeling comprehension was assessed by HCP ratings of subject proficiency. Key system features and ease-of-use were evaluated by subject questionnaires. Results All subjects who completed the study (N = 74) successfully performed blood glucose measurements, connected the meter to a laptop computer, and used key features of the system. The system was accurate; 98.6% (146/148) of subject results and 96.6% (143/148) of HCP results exceeded ISO 15197:2003 criteria. All subject and HCP results were clinically accurate (97.3%; zone A) or associated with benign errors (2.7%; zone B). The majority of subjects rated features of the BGMS as “very good” or “excellent.” Conclusions CONTOUR USB exceeded ISO 15197:2003 system performance criteria in the hands of untrained lay users. Subjects understood the product labeling, found the system easy to use, and successfully performed blood glucose testing. PMID:22027308

  7. Blood glucose self-monitoring among patients with diabetes mellitus type 2 in family medicine practice

    Directory of Open Access Journals (Sweden)

    Herenda Samira

    2009-01-01

    Full Text Available Introduction. Good knowledge of diabetic patients about their disease is often not related with good glycemic control. The aim of this study was to determine the level of application of acquired knowledge about diabetes in recognizing and resolving hypoglycemic and hyperglycemic conditions in patients who did or did not do blood glucose self-monitoring as well as the impact of self-monitoring on HbA1c during education of patients with diabetes type 2. Material and methods. There were 91 patients with the type 2 diabetes who completed six months education about their disease in four family medicine practices in Tuzla Canton during the period from March to September 2005. The patients who did or did not do self-monitoring with glucometer were interviewed on knowledge about recognizing and resolving hypoglycemia and hyperglycemia by family physician and HbA1c was assessed at the beginning of the education, 3 months after reading the brochure (passive education and additional 3 months of group (intensive education. Results. Out of 91 interviewed patients, there were 29 who did self-monitoring by glucometer at the beginning of the education, 30 patients during the passive education and 34 after the intensive education. At the beginning of education, regardless of doing self-monitoring, the patients were less able to recognize and resolve hypoglycemia and even less hyperglycemia. At the end of education, their knowledge was better at both recognizing and resolving hypoglycemia (P=0.01 as well as at recognizing (P=0.01 and resolving hyperglycemia (P=0.001. In the patients who did self-monitoring the average value of the HbA1c did not improve significantly (P=0.44 compared to those who did not practice self-monitoring (P=0.10 during education. Conclusion. Only one third of patients with type 2 diabetes had done self-monitoring with glucometer and although their knowledge about hypoglycemia and hyperglycemia was improved during education, these patients

  8. Portable Mass Spectrometer System for in-situ Environmental Gas Monitoring

    Science.gov (United States)

    Conejo, E.; Griffin, T. P.; Diaz, J. A.; Arkin, C. R.; Soto, C.; Naylor, G. R.; Curley, C.; Floyd, D.

    2005-01-01

    A system developed by NASA has been used for monitoring air quality around different locations. The system was designed for aircraft applications but has proven to be very useful as a portable gas analyzer. The system has been used to monitor air quality around volcanoes, cities, and the surrounding areas. The transport of the system has been via aircraft, car, and hand carried.

  9. Transcutaneous Spinal Direct Current Stimulation (tsDCS

    Directory of Open Access Journals (Sweden)

    Filippo eCogiamanian

    2012-07-01

    Full Text Available In the past ten years renewed interest has centered on non-invasive transcutaneous weak direct currents applied over the scalp to modulate cortical excitability (brain polarization or transcranial direct current stimulation, tDCS. Extensive literature shows that tDCS induces marked changes in cortical excitability that outlast stimulation.Aiming at developing a new, non invasive, approach to spinal cord neuromodulation we assessed the after-effects of thoracic transcutaneous spinal DC stimulation (tsDCS on somatosensory potentials (SEPs evoked in healthy subjects by posterior tibial nerve (PTN stimulation. Our findings showed that thoracic anodal tsDCS depresses the cervico-medullary PTN-SEP component (P30 without eliciting adverse effects. tsDCS also modulates post-activation H-reflex dynamics. Later works further confirmed that transcutaneous electric fields modulate spinal cord function. Subsequent studies in our laboratory showed that tsDCS modulates the flexion reflex in the human lower limb. Besides influencing the laser evoked potentials, tsDCS increases pain tolerance in healthy subjects. Hence, though the underlying mechanisms remain speculative, tsDCS modulates activity in lemniscal, spinothalamic and segmental motor systems.Here we review currently available experimental evidence that non-invasive spinal cord stimulation influences spinal function in humans and argue that, by focally modulating spinal excitability, tsDCS could provide a novel therapeutic tool complementary to drugs and invasive spinal cord stimulation in managing various pathologic conditions, including pain.

  10. Transcutaneous mechanical nerve stimulation using perineal vibration: a novel method for the treatment of female stress urinary incontinence

    DEFF Research Database (Denmark)

    Sønksen, Jens; Ohl, Dana A; Bonde, Birthe;

    2007-01-01

    We defined basic guidelines for transcutaneous mechanical nerve stimulation in modifying pelvic floor responses in women and determined the efficacy of transcutaneous mechanical nerve stimulation in treating stress urinary incontinence....

  11. Effect of training on the performance of blood glucose monitoring using a reagent strip (Glucoprofil).

    Science.gov (United States)

    Chantelau, E; Starostina, E G; Antsiferov, M B

    1990-11-01

    A relatively simple, new test-strip for blood-glucose monitoring, the Glucoprofil strip with a reaction-zone of film-foil, was evaluated by testing 100 different blood glucose concentrations between 2.1 and 29.7 mmol/l. The results were obtained visually before and after training to read the color changes of the strip according to the scale on the strip-container. Plasma glucose measurements with the Beckman Glucose-Analyzer were used for reference. The results show that the Glucoprofil-strip readings correlated well with the reference method (r = 0.97), thorough training provided. The mean deviation of the strip results was less than 1 mmol/l. Similar results were obtained using another blood glucose strip, the Haemoglukotest 20-800 R. Our study indicates that the performance of the Glucoprofil strip is satisfactory, and hence the strip may be useful for clinical purposes. Consistent with previous reports we could demonstrate that training improved the reading results of the Glucoprofil strip.

  12. System Accuracy Evaluation of the GlucoRx Nexus Voice TD-4280 Blood Glucose Monitoring System

    Directory of Open Access Journals (Sweden)

    Muhammad Khan

    2014-01-01

    Full Text Available Use of blood glucose (BG meters in the self-monitoring of blood glucose (SMBG significantly lowers the risk of diabetic complications. With several BG meters now commercially available, the International Organization for Standardization (ISO ensures that each BG meter conforms to a set degree of accuracy. Although adherence to ISO guidelines is a prerequisite for commercialization in Europe, several BG meters claim to meet the ISO guidelines yet fail to do so on internal validation. We conducted a study to determine whether the accuracy of the GlucoRx Nexus TD-4280 meter, utilized by our department for its cost-effectiveness, complied with ISO guidelines. 105 patients requiring laboratory blood glucose analysis were randomly selected and reference measurements were determined by the UniCel DxC 800 clinical system. Overall the BG meter failed to adhere to the ≥95% accuracy criterion required by both the 15197:2003 (overall accuracy 92.4% and 15197:2013 protocol (overall accuracy 86.7%. Inaccurate meters have an inherent risk of over- and/or underestimating the true BG concentration, thereby risking patients to incorrect therapeutic interventions. Our study demonstrates the importance of internally validating the accuracy of BG meters to ensure that its accuracy is accepted by standardized guidelines.

  13. Ambulatory blood pressure monitoring-derived short-term blood pressure variability is increased in Cushing's syndrome.

    Science.gov (United States)

    Rebellato, Andrea; Grillo, Andrea; Dassie, Francesca; Sonino, Nicoletta; Maffei, Pietro; Martini, Chiara; Paoletta, Agostino; Fabris, Bruno; Carretta, Renzo; Fallo, Francesco

    2014-11-01

    Cushing's syndrome is associated with high cardiovascular morbility and mortality. Blood pressure (BP) variability within a 24-h period is increasingly recognized as an independent predictor of cardiovascular risk. The aim of our study was to investigate the short-term BP variability indices in Cushing's syndrome. Twenty-five patients with Cushing's syndrome (mean age 49 ± 13 years, 4 males; 21 Cushing's disease and 4 adrenal adenoma patients) underwent 24-h ambulatory BP monitoring (ABPM) and evaluation of cardiovascular risk factors. Cushing patients were divided into 8 normotensive (NOR-CUSH) and 17 hypertensive (HYP-CUSH) patients and were compared with 20 normotensive (NOR-CTR) and 20 hypertensive (HYP-CTR) age-, sex-, and BMI-matched control subjects. Short-term BP variability was derived from ABPM and calculated as the following: (1) standard deviation (SD) of 24-h, daytime, and nighttime BP; (2) 24-h weighted SD of BP; and (3) average real variability (ARV), i.e., the average of the absolute differences between consecutive BP measurements over 24 h. In comparison with controls, patients with Cushing's syndrome, either normotensive or hypertensive, had higher 24-h and daytime SD of BP, as well as higher 24-h weighted SD and ARV of BP (P = 0.03 to P CUSH or between HYP-CTR and HYP-CUSH subgroups. ABPM-derived short-term BP variability is increased in Cushing's syndrome, independent of BP elevation. It may represent an additional cardiovascular risk factor in this disease. The role of excess cortisol in BP variability has to be further clarified.

  14. Development of non-contact beam intensity monitor by residual gas ionization

    International Nuclear Information System (INIS)

    Non-contact beam intensity monitor is an important device for high-intensity proton accelerator, such as J-PARC. Especially, it is difficult to measure beam intensity of DC-like beam with off-contact. We have developed non-contact beam intensity monitor for J-PARC slow-extraction beam line by measuring charge of ionization electrons in 1 Pa residual gas. In this article, specifications of the residual gas ionization current monitor (RGICM) and results of a test experiment with 65 MeV proton beam at Research Center for Nuclear Physics in Osaka University are reported. (author)

  15. Arterial blood gas analysis and electrolyte determination in neonates with asphyxia

    Institute of Scientific and Technical Information of China (English)

    Zi-Mei Sun

    2015-01-01

    Objective:To explore the value of arterial blood gas analysis and electrolyte determination in the diagnosis of neonatal asphyxia.Methods: A total of 100 neonates with asphyxia who were admitted in our department from March, 2013 to March, 2014 were included in the study and divided into the mild asphyxia group and the severe asphyxia group according to Apgar scoring. Moreover, 50 normal neonates were served as the control group and used for comparative analysis. AVL blood-gas analyzer was used to detect the levels of pH, PaO2, HCO3-, BE, PaCO2, K+, Na+, Cl-, and Ca2+.Results:The concentrations of pH, PaO2, HCO3-, and BE in the severe asphyxia group were significantly lower than those in the mild asphyxia group and the control group, while PaCO2 level was significantly higher than that in the mild asphyxia group and the control group. The comparison of the various indicators between the control group and the mild asphyxia group was not statistically significant. The comparison of serum K+, Na+,and Cl- levels among the control group, the mild asphyxia group, and the severe asphyxia group was not statistically significant. The serum Ca2+ level in the severe asphyxia group was significantly lower than that in the mild asphyxia group and the control group, while the comparison between the mild asphyxia group and the control group was also statistically significant.Conclusions:Blood gas analysis and electrolyte determination to the arterial blood in neonates can compensate for the insufficiency of Apgar scoring and provide an objective evidence for the diagnosis of neonatal asphyxia and the estimation of severity degree. Clinical combination with Apgar scoring is of great significance in enhancing the diagnosis of neonatal asphyxia and the accuracy of severity evaluation.

  16. Transcutaneous oxygen tension in imminent foot gangrene

    DEFF Research Database (Denmark)

    Tønnesen, K H

    1978-01-01

    with moderate obstruction, the oxygen tension reached zero at a toe systolic blood pressure of 5--10 mmHg (tilt toe up) and reached arterial oxygen tension at about 50 to 70 mmHg (tilt toe down). In legs withsevere arterial obstruction and ischaemic rest pain, oxygen tension rose from zero not before systolic...

  17. Fast gas chromatography and negative-ion chemical ionization tandem mass spectrometry for forensic analysis of cannabinoids in whole blood.

    Science.gov (United States)

    Thomas, Aurélien; Widmer, Christèle; Hopfgartner, Gérard; Staub, Christian

    2007-11-01

    The present work describes a fast gas chromatography/negative-ion chemical ionization tandem mass spectrometric assay (Fast GC/NICI-MS/MS) for analysis of tetrahydrocannabinol (THC), 11-hydroxy-tetrahydrocannabinol (THC-OH) and 11-nor-9-carboxy-tetrahydrocannabinol (THC-COOH) in whole blood. The cannabinoids were extracted from 500 microL of whole blood by a simple liquid-liquid extraction (LLE) and then derivatized by using trifluoroacetic anhydride (TFAA) and hexafluoro-2-propanol (HFIP) as fluorinated agents. Mass spectrometric detection of the analytes was performed in the selected reaction-monitoring mode on a triple quadrupole instrument after negative-ion chemical ionization. The assay was found to be linear in the concentration range of 0.5-20 ng/mL for THC and THC-OH, and of 2.5-100 ng/mL for THC-COOH. Repeatability and intermediate precision were found less than 12% for all concentrations tested. Under standard chromatographic conditions, the run cycle time would have been 15 min. By using fast conditions of separation, the assay analysis time has been reduced to 5 min, without compromising the chromatographic resolution. Finally, a simple approach for estimating the uncertainty measurement is presented. PMID:17913432

  18. Non-constrained monitoring of systolic blood pressure on a weighing scale

    International Nuclear Information System (INIS)

    In this study, we developed a novel technique for estimating non-constrained and cuffless blood pressure (BP) that was based on electrocardiogram (ECG) and ballistocardiogram (BCG). The BCG was non-invasively measured using a common electronic weighing scale when a subject was standing on it. The ECG was measured using three different methods: on the chest using Ag/AgCl electrodes, on the hands using dry electrodes and on the feet also using dry electrodes. For a BP correlated parameter, a time interval parameter, which was defined as the time difference between the ECG R-peak and BCG J-peak, was employed for evaluating and estimating beat-to-beat BP. Under a BP varying experiment with a Valsalva manoeuvre, the R–J intervals were extracted at every beat cycle and a systolic blood pressure (SBP) estimation equation was established using linear regression analysis for each subject. In the case of feet delivered ECG (F-ECG), an ensemble average technique synchronized at the BCG J-peak point was applied to extract the ECG signal from the feet. The performance of the proposed method was evaluated using Finapres, a non-invasive blood pressure measurement system, as a reference BP signal, and a scatter plot was used to find the regression line between the reference values and estimated BPs. A moving-window averaging technique was applied to remove the high-frequency noise in the R–J intervals and was applied to enhance the accuracy of the SBP estimation. For all individuals, the estimated SBP was similar to the measured SBP with a reliable correlation, which makes the proposed method suitable for use in a home healthcare system to monitor blood pressure on a weighing scale at the same time as measuring weight

  19. Review of Monitoring Plans for Gas Bubble Disease Signs and Gas Supersaturation Levels on the Columbia and Snake Rivers.

    Energy Technology Data Exchange (ETDEWEB)

    Fidler, Larry; Elston, Ralph; Colt, John

    1994-07-01

    Montgomery Watson was retained by the Bonneville Power Administration to evaluate the monitoring program for gas bubble disease signs and dissolved gas supersaturation levels on the Columbia and Snake rivers. The results of this evaluation will provide the basis for improving protocols and procedures for future monitoring efforts. Key study team members were Dr. John Colt, Dr. Larry Fidler, and Dr. Ralph Elston. On the week of June 6 through 10, 1994 the study team visited eight monitoring sites (smolt, adult, and resident fish) on the Columbia and Snake rivers. Additional protocol evaluations were conducted at the Willard Field Station (National Biological Survey) and Pacific Northwest Laboratories at Richland (Battelle). On June 13 and 14, 1994, the study team visited the North Pacific Division office of the U.S. Corps of Engineers and the Fish Passage Center to collect additional information and data on the monitoring programs. Considering the speed at which the Gas Bubble Trauma Monitoring Program was implemented this year, the Fish Passage Center and cooperating Federal, State, and Tribal Agencies have been doing an incredible job. Thirty-one specific recommendations are presented in this report and are summarized in Section 14.

  20. Diabetes Technology: Markers, Monitoring, Assessment, and Control of Blood Glucose Fluctuations in Diabetes

    Directory of Open Access Journals (Sweden)

    Boris P. Kovatchev

    2012-01-01

    Full Text Available People with diabetes face a life-long optimization problem: to maintain strict glycemic control without increasing their risk for hypoglycemia. Since the discovery of insulin in 1921, the external regulation of diabetes by engineering means has became a hallmark of this optimization. Diabetes technology has progressed remarkably over the past 50 years—a progress that includes the development of markers for diabetes control, sophisticated monitoring techniques, mathematical models, assessment procedures, and control algorithms. Continuous glucose monitoring (CGM was introduced in 1999 and has evolved from means for retroactive review of blood glucose profiles to versatile reliable devices, which monitor the course of glucose fluctuations in real time and provide interactive feedback to the patient. Technology integrating CGM with insulin pumps is now available, opening the field for automated closed-loop control, known as the artificial pancreas. Following a number of in-clinic trials, the quest for a wearable ambulatory artificial pancreas is under way, with a first prototype tested in outpatient setting during the past year. This paper discusses key milestones of diabetes technology development, focusing on the progress in the past 10 years and on the artificial pancreas—still not a cure, but arguably the most promising treatment of diabetes to date.

  1. Gas dynamics considerations in a non-invasive profile monitor for charged particle beams

    CERN Document Server

    Tzoganis, Vasilis; Welsch, Carsten P

    2014-01-01

    A non-invasive, gas jet-based, beam profile monitor has been developed in the QUASAR Group at the Cockcroft Institute, UK. This allows on-line measurement of the 2-dimensional transverse profile of particle beams with negligible disturbance to either primary beam or accelerator vacuum. The monitor is suitable for use with beams across a wide range of energies and intensities. In this setup a nozzle-skimmer system shapes a thin supersonic gas jet into a curtain. However, the small dimensions of the gas inlet nozzle and subsequent skimmers were shown to be the cause of many operational problems. In this paper, the dynamics of gas jet formation transport and shaping is discussed before an image-processing based alignment technique is introduced. Furthermore, experimental results obtained with a 5 keV electron beam are discussed and the effects of gas stagnation pressure on the acquired beam are presented.

  2. Gas monitoring data anomaly identification based on spatio-temporal correlativity analysis

    Institute of Scientific and Technical Information of China (English)

    Shi-song ZHU; Yun-jia WANG; Lian-jiang WEI

    2013-01-01

    Based on spatio-temporal correlativity analysis method,the automatic identification techniques for data anomaly monitoring of coal mining working face gas are presented.The asynchronous correlative characteristics of gas migration in working face airflow direction are qualitatively analyzed.The calculation method of asynchronous correlation delay step and the prediction and inversion formulas of gas concentration changing with time and space after gas emission in the air return roadway are provided.By calculating one hundred and fifty groups of gas sensors data series from a coal mine which have the theoretical correlativity,the correlative coefficient values range of eight kinds of data anomaly is obtained.Then the gas monitoring data anomaly identification algorithm based on spatio-temporal correlativity analysis is accordingly presented.In order to improve the efficiency of analysis,the gas sensors code rules which can express the spatial topological relations are suggested.The experiments indicate that methods presented in this article can effectively compensate the defects of methods based on a single gas sensor monitoring data.

  3. Validation protocol for multiple blood gas analyzers in accordance with laboratory accreditation programs

    Directory of Open Access Journals (Sweden)

    Pérsio A. R. Ebner

    2015-10-01

    Full Text Available ABSTRACTIntroduction:The results of blood gas analysis using different instrumentation can vary widely due to the methodological differences, the calibration procedures and the use of different configurations for each type of instrument.Objective:The objective of this study was to evaluate multiple analytical systems for measurement of blood gases, electrolytes and metabolites in accordance with the accreditation program (PALC of Sociedade Brasileira de Patologia Clínica/Medicina Laboratorial (SBPC/ML.Materials and methods:20 samples were evaluated in three ABL800 Flex (Radiometer Medical ApS, Denmark blood gas analyzers, and the results were compared with those of the device in use, which was considered the reference. The analysis of variance (Anova was applied for statistical purposes, as well as the calculation of mean, standard deviation and coefficient of variation.Results:The p values obtained in the statistical analysis were: pH = 0.983, pO2 = 0.991, pCO2 = 0.353, lactate = 0.584, glucose = 0.995, ionized calcium = 0.983, sodium = 0.991, potassium = 0.926, chlorine = 0.029.Conclusion:The evaluation of multiple analytical systems is an essential procedure in the clinical laboratory for quality assurance and accuracy of the results.

  4. PL 03-2 REMOTE BLOOD PRESSURE MONITORING IN HYPERTENSION CARE: PRACTICE OF HYPERTENSION BY TELETRANSMISSION OF HOME BLOOD PRESSURE.

    Science.gov (United States)

    Imai, Yutaka

    2016-09-01

    It has been confirmed that clinical significance of home blood pressure measurements (HBPM) is higher than clinic BP measurements and ambulatory BP monitoring. However, several drawbacks of HBPM have also been mentioned, e.g. selection and reporting biases, difficulties of calculation of multiple measurements, difficulties of onsite judgement of numerous recordings, etc. Recent devices for HBPM incorporate memory function. This function can overcome such drawbacks of HBPM. These memorized data can transmit, storage, retrieve, be arithmetic and control, be judged based on algorithm and be got feedback. These functions are available for telemonitoring and telemedicine. The purpose of telemedicine based on HBPM is as a tool for proper diagnosis and treatment of hypertension, as a supporting system for practice of hypertension through rapid and proper feedback and as a tool of self-medication. Telemonitoring using HBPM is useful for clinical pharmacology of antihypertensive drugs and can be applied to large scale intervention trial of antihypertensive drugs such as THOP Study and HOMED-BP study. We have successfully conducted HOME-BP study, a randomized controlled study, based on HBPM and interactive system using Internet, where 3500 hypertensive subjects were followed for maximal 10 years by 300 general practitioners. HOMED-BP study indicated that telemedicine has been practiced without trouble and accepted readily by practitioners and patients. In general, it has been demonstrated that patients show a high degree of adherence to telemonitoring program and regular use of these technologies in predisposition of hypertensive patients to use electronic BP monitors at home. However, older patients may express difficulty in using the HBP transmitting system.Home telemedicine of HBPM brings correct and reliable data during HOMED-BP study and HOMED-BP study provided goal HBP level for mild to moderate hypertensive patients; ie, 130 mmHg of systolic HBP. New telemedicine

  5. Portable Mass Spectrometer Applications for In Situ Environmental Gas Monitoring

    Science.gov (United States)

    Griffin, Timothy P.; Diaz, J. Andres; Arkin, C. Richard; Conejo, Elian

    2005-01-01

    Primary Goal of this project is to (1) Design/build a flexible system to monitor air contamination (2) Learn requirements for operating system in low pressure and low temperature environments (3) Design/build system for integration into aircraft and automobiles Secondary Goals/Offshoots are (1) Fly aboard different aircraft (2)Hand-carry unit (3) Drive unit in automobiles.

  6. Dried blood spots for viral load monitoring in Malawi: feasible and effective.

    Directory of Open Access Journals (Sweden)

    Sarah E Rutstein

    Full Text Available To evaluate the feasibility and effectiveness of dried blood spots (DBS use for viral load (VL monitoring, describing patient outcomes and programmatic challenges that are relevant for DBS implementation in sub-Saharan Africa.We recruited adult antiretroviral therapy (ART patients from five district hospitals in Malawi. Eligibility reflected anticipated Ministry of Health VL monitoring criteria. Testing was conducted at a central laboratory. Virological failure was defined as >5000 copies/ml. Primary outcomes were program feasibility (timely result availability and patient receipt and effectiveness (second-line therapy initiation.We enrolled 1,498 participants; 5.9% were failing at baseline. Median time from enrollment to receipt of results was 42 days; 79.6% of participants received results within 3 months. Among participants with confirmed elevated VL, 92.6% initiated second-line therapy; 90.7% were switched within 365 days of VL testing. Nearly one-third (30.8% of participants with elevated baseline VL had suppressed (4 years were more likely to be failing than participants on therapy 1-4 years (RR 1.7, 95% CI 1.0-2.8; older participants were less likely to be failing (RR 0.95, 95% CI 0.92-0.98. There was no difference in likelihood of failure based on clinical symptoms (RR 1.17, 95% CI 0.65-2.11.DBS for VL monitoring is feasible and effective in real-world clinical settings. Centralized DBS testing may increase access to VL monitoring in remote settings. Programmatic outcomes are encouraging, especially proportion of eligible participants switched to second-line therapy.

  7. The future of gas turbine compliance monitoring: The integration of PEMS and CEMS for regulatory compliance

    Energy Technology Data Exchange (ETDEWEB)

    Macak, J.J. III

    1999-07-01

    When the New Source Performance Standards (NSPS) for Stationary Gas Turbines were first promulgated in 1979 (40 CFR 60, Subpart GG), continuous compliance monitoring for gas turbines was simply a parametric monitoring approach where a unit was tested at four load conditions. For those units where water or steam injection was used for NO{sub x} control, testing consisted of establishing a water (or steam injection) versus fuel flow curve to achieve permitted NO{sub x} emission levels across the load range. Since 1979, the growth in gas turbine popularity has encouraged the development of Predictive Emissions Monitoring Systems (PEMS) where gas turbine operating parameters and ambient conditions are fed into a prediction algorithm to predict, rather than monitor, emissions. However, permitting requirements and technological advances now have gas turbines emitting NO{sub x} in the single digits while the overall combined-cycle thermal efficiency has improved dramatically. The combination of supplemental duct-firing in heat recovery steam generators, pollution prevention technology, post-combustion emission controls, and EPA Continuous Emissions Monitoring System (CEMS) regulations for the power industry, resulted in a shift towards CEMS due to the complexity of the overall process. Yet, CEMS are often considered to be a maintenance nightmare with significant amounts of downtime. CEMS and PEMS have their own advantages and disadvantages. Thus evolved the need to find the optimum balance between CEMS and PEMS for gas turbine projects. To justify the cost of both PEMS and CEMS in the same installation, there must be an economic incentive to do so. This paper presents the application of a PEMS/CEMS monitoring system that integrates both PEMS and CEMS in order to meet, and exceed, all emissions monitoring requirements.

  8. Is liquid heparin comparable to dry balanced heparin for blood gas sampling in intensive care unit?

    Directory of Open Access Journals (Sweden)

    Viswas Chhapola

    2014-01-01

    Full Text Available Introduction: Blood gas (BG analysis is required for management of critically ill patients in emergency and intensive care units. BG parameters can be affected by the type of heparin formulations used-liquid heparin (LH or dry balanced heparin (DBH. This study was conducted to determine whether blood gas, electrolyte, and metabolite estimations performed by using DBH and LH are comparable. Materials and Methods: A prospective study was conducted at pediatric intensive care unit (PICU of a tertiary care hospital. Paired venous samples were collected from 35 consecutive children in commercially prepared DBH syringes and custom-prepared LH syringes. Samples were immediately analyzed by blood gas analyzer and compared for pH, pCO 2 , pO 2 , HCO 3 - , Na + , K + , Cl - , and lactate. Paired comparisons were done and agreement was assessed by Bland-Altman difference plots. The 95% limits of absolute agreement (LOA were compared with the specifications for total allowable error (TEa. Results: The P values were significant for all measured parameters, with the exception of pCO 2 and K +. Bland-Altman difference plots showed wide LOA for pCO 2 , pO 2 , HCO3 - , Na + , K + , and Cl - when compared against TEa. For pCO 2 , HCO3 - , Na + , K + , and Cl - , 40%, 23%, 77%, 34%, and 54% of samples were outside the TEa limits, respectively, with LH. Conclusion: Our study showed that there is poor agreement between LH and DBH for the BG parameters pCO2, pO2, HCO3 - , K + , Na + , and Cl - and, thus, are not comparable. But for pH and lactate, LH and DBH can be used interchangeably.

  9. The usefulness of the arterial blood gas in pure carbon monoxide poisoning.

    Science.gov (United States)

    Lebby, T I; Zalenski, R; Hryhorczuk, D O; Leikin, J B

    1989-04-01

    In a retrospective study of 49 cases of carbon monoxide (CO) intoxication presented to the University of Illinois Hospital (UIH) Emergency Department between November 1986 and April 1988, we looked for a correlation between carboxyhemoglobin (COHb) as determined by a venous sample and the pH as determined by arterial blood gas analysis. The range of COHb levels in our study was 10-64% (mean 21.8% +/- 10.2%). Smoke inhalation cases (n = 3) were excluded from our study because they did not represent pure CO intoxication. Of the remaining 46 cases, 18 had arterial blood gases drawn. In none of these 18 cases (mean COHb 24.5% +/- 12.6%) did we find a correlation between COHb levels and the pH as determined by linear regression analysis. Also, in none of the 18 cases were there any therapeutic interventions associated with the arterial blood gas result. Additionally, in none of the remaining 28 cases were any therapeutic interventions performed with regards to patients' acidosis or ventilatory status (except 100% oxygen administration. We also retrospectively reviewed records of 104 cases who presented to Cook County Hospital Emergency Department with COHb levels over 10% during the period between March 1986 and May 1988. In these cases, we found no significant correlation between COHb level and arterial pH. We therefore conclude that arterial blood gases drawn in order to determine the degree of acidosis in mild CO intoxication without respiratory distress may not be useful in guiding therapeutic intervention and need not be routinely drawn. PMID:2929122

  10. Environmental Monitoring and the Gas Industry: Program Manager Handbook

    Energy Technology Data Exchange (ETDEWEB)

    Gregory D. Gillispie

    1997-12-01

    This document has been developed for the nontechnical gas industry manager who has the responsibility for the development of waste or potentially contaminated soil and groundwater data or must make decisions based on such data for the management or remediation of these materials. It explores the pse of common analytical chemistry instrumentation and associated techniques for identification of environmentally hazardous materials. Sufficient detail is given to familiarize the nontechnical reader with the principles behind the operation of each technique. The scope and realm of the techniques and their constituent variations are portrayed through a discussion of crucial details and, where appropriate, the depiction of real-life data. It is the author's intention to provide an easily understood handbook for gas industry management. Techniques which determine the presence, composition, and quantification of gas industry wastes are discussed. Greater focus is given to traditional techniques which have been the mainstay of modem analytical benchwork. However, with the continual advancement of instrumental principles and design, several techniques have been included which are likely to receive greater attention in fiture considerations for waste-related detection. Definitions and concepts inherent to a thorough understanding of the principles common to analytical chemistry are discussed. It is also crucial that gas industry managers understand the effects of the various actions which take place before, during, and after the actual sampling step. When a series of sample collection, storage, and transport activities occur, new or inexperienced project managers may overlook or misunderstand the importance of the sequence. Each step has an impact on the final results of the measurement process; errors in judgment or decision making can be costly. Specific techniques and methodologies for the collection, storage, and transport of environmental media samples are not

  11. An in situ method for real-time monitoring of soil gas diffusivity

    Science.gov (United States)

    Laemmel, Thomas; Maier, Martin; Schack-Kirchner, Helmer; Lang, Friederike

    2016-04-01

    Soil aeration is an important factor for the biogeochemistry of soils. Generally, gas exchange between soil and atmosphere is assumed to be governed by molecular diffusion and by this way fluxes can be calculated using by Fick's Law. The soil gas diffusion coefficient DS represents the proportional factor between the gas flux and the gas concentration gradient in the soil and reflects the ability of the soil to "transport passively" gas through the soil. One common way to determine DS is taking core samples in the field and measuring DS in the lab. Unfortunately this method is destructive and laborious and it can only reflect a small fraction of the whole soil. As a consequence, uncertainty about the resulting effective diffusivity on the profile scale, i.e. the real aeration status remains. We developed a method to measure and monitor DS in situ. The set-up consists of a custom made gas sampling device, the continuous injection of an inert tracer gas and inverse gas transport modelling in the soil. The gas sampling device has seven sampling depths (from 0 to -43 cm of depth) and can be easily installed into vertical holes drilled by an auger, which allows for fast installation of the system. Helium (He) as inert tracer gas was injected continuously at the lower end of the device. The resulting steady state distribution of He was used to deduce the DS depth distribution of the soil. For Finite Element Modeling of the gas-sampling-device/soil system the program COMSOL was used. We tested our new method both in the lab and in a field study and compared the results with a reference lab method using soil cores. DS profiles obtained by our in-situ method were consistent with DS profiles determined based on soil core analyses. Soil gas profiles could be measured with a temporal resolution of 30 minutes. During the field study, there was an important rain event and we could monitor the decrease in soil gas diffusivity in the top soil due to water infiltration. The effect

  12. Determination of ketone bodies in blood by headspace gas chromatography-mass spectrometry

    DEFF Research Database (Denmark)

    Holm, Karen Marie Dollerup; Linnet, Kristian; Rasmussen, Brian Schou;

    2010-01-01

    A gas chromatography-mass spectrometry (GC-MS) method for determination of ketone bodies (ß-hydroxybutyrate, acetone, and acetoacetate) in blood is presented. The method is based on enzymatic oxidation of D-ß-hydroxybutyrate to acetoacetate, followed by decarboxylation to acetone, which...... was quantified by the use of headspace GC-MS using acetone-(13)C(3) as an internal standard. The developed method was found to have intra- and total interday relative standard deviations acetone+acetoacetate levels (~25 to 8300 µM) and D-ß-hydroxybutyrate levels (~30 to 16500 µM). Recovery values varied...

  13. A rapid and reversible skull optical clearing method for monitoring cortical blood flow

    Science.gov (United States)

    Zhang, Chao; Zhao, Yanjie; Shi, Rui; Zhu, Dan

    2016-03-01

    In vivo cortex optical imaging is of great important for revealing both structural and functional architecture of brain with high temporal-spatial resolution. To reduce the limitation of turbid skull, researchers had to establish various skull windows or directly expose cortex through craniotomy. Here we developed a skull optical clearing method to make skull transparent. Laser speckle contrast imaging technique was used to monitor the cortical blood flow after topical treatment with the optical clearing agents. The results indicated that the image contrast increased gradually, and then maintained at a high level after 15 min for adult mice, which made the image quality and resolution of micro-vessels nearly approximate to those of exposed cortex. Both the cortical blood flow velocity almost kept constant after skull became transparent. Besides, the treatment of physiological saline on the skull could make skull return to the initial state again and the skull could become transparent again when SOCS retreated it. Thus, we could conclude that the skull optical clearing method was rapid, valid, reversible and safe, which provided us available approach for performing the cortical structural and functional imaging at high temporal-spatial resolution.

  14. Fully Autonomous Multiplet Event Detection: Application to Local-Distance Monitoring of Blood Falls Seismicity

    Energy Technology Data Exchange (ETDEWEB)

    Carmichael, Joshua Daniel [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Carr, Christina [Univ. of Alaska, Fairbanks, AK (United States); Pettit, Erin C. [Univ. of Alaska, Fairbanks, AK (United States)

    2015-06-18

    We apply a fully autonomous icequake detection methodology to a single day of high-sample rate (200 Hz) seismic network data recorded from the terminus of Taylor Glacier, ANT that temporally coincided with a brine release episode near Blood Falls (May 13, 2014). We demonstrate a statistically validated procedure to assemble waveforms triggered by icequakes into populations of clusters linked by intra-event waveform similarity. Our processing methodology implements a noise-adaptive power detector coupled with a complete-linkage clustering algorithm and noise-adaptive correlation detector. This detector-chain reveals a population of 20 multiplet sequences that includes ~150 icequakes and produces zero false alarms on the concurrent, diurnally variable noise. Our results are very promising for identifying changes in background seismicity associated with the presence or absence of brine release episodes. We thereby suggest that our methodology could be applied to longer time periods to establish a brine-release monitoring program for Blood Falls that is based on icequake detections.

  15. Microvascular blood flow monitoring with laser speckle contrast imaging using the generalized differences algorithm.

    Science.gov (United States)

    Humeau-Heurtier, Anne; Mahé, Guillaume; Abraham, Pierre

    2015-03-01

    Laser speckle contrast imaging (LSCI) is a full-field optical technique to monitor microvascular blood flow with high spatial and temporal resolutions. It is used in many medical fields such as dermatology, vascular medicine, or neurosciences. However, LSCI leads to a large amount of data: image sampling frequency is often of several Hz and recordings usually last several minutes. Therefore, clinicians often perform regions of interest in which a spatial averaging of blood flow is performed and the result is followed with time. Unfortunately, this leads to a poor spatial resolution for the analyzed data. At the same time, a higher spatial resolution for the perfusion maps is wanted. To get over this dilemma we propose a new post-acquisition visual representation for LSCI perfusion data using the so-called generalized differences (GD) algorithm. From a stack of perfusion images, the procedure leads to a new single image with the same spatial resolution as the original images and this new image reflects perfusion changes. The algorithm is herein applied on simulated stacks of images and on experimental LSCI perfusion data acquired in three different situations with a commercialized laser speckle contrast imager. The results show that the GD algorithm provides a new way of visualizing LSCI perfusion data. PMID:25576743

  16. Effects of SO2 and pH on blood-gas partition coefficients of inert gases.

    Science.gov (United States)

    Yamaguchi, K; Mori, M; Kawai, A; Asano, K; Takasugi, T; Umeda, A; Yokoyama, T

    1990-01-01

    Potential effects of SO2 and of pH on blood-gas partition coefficients, lambda, for inert gases, including SF6, ethane, cyclopropane, halothane, diethyl ether, acetone and N2, were systematically investigated using human blood. Measurements on lambda were performed at 37 degrees C in conditions of varied SO2 and pH using gas chromatography. Incorporating the experimental data on lambda, multiple inert gas elimination was applied to 18 patients with varied chronic lung diseases, in order to estimate the effects of SO2 and of pH on both inert gas exchange and resultant recovery of VA/Q distribution in the lung. For this purpose, the data obtained by the procedure of multiple inert gas elimination were analyzed with the classical approach but allowance was made for lambda of the indicator gas to vary according to exchange of O2 and of CO2 in the pulmonary capillary. Among the gases studied, ethane, cyclopropane, halothane and diethyl ether showed significantly smaller lambda values in the oxygenated blood than in deoxygenated blood, whereas SF6, acetone and N2 were little dependent on SO2. An increase in lambda was found for ethane and a decrease for halothane with increasing pH in the blood. The other gases were not significantly influenced by pH. In spite of these experimental findings, regional difference of either SO2 or pH in the lung did not exert important influence on the inert gas exchange or on the predicted VA/Q distribution. In conclusion, blood-gas partition coefficients of some inert gases are consistently altered by SO2 and pH, but their possible effects on inert gas exchange seem to be negligible. PMID:1965757

  17. Validation of pulse oximetry for monitoring of hypoxaemic episodes in the late postoperative period

    DEFF Research Database (Denmark)

    Stausholm, K; Rosenberg-Adamsen, S; Edvardsen, L;

    1997-01-01

    We monitored 12 patients undergoing major abdominal surgery using a pulse oximeter (Nellcor N-200) and a transcutaneous oxygen tension monitor (TINA, Radiometer A/S) on the second or third night after operation. Of the shortest hypoxaemic episodes measured with the pulse oximeter (... duration), 78% also occurred in the transcutaneous oxygen tension measurement. Episodes of longer duration (> or = 1 min duration on the pulse oximeter) were, in 95% of cases, reflected in the transcutaneous oxygen tension measurement also. Thus postoperative episodic desaturations lasting > or = 1 min...

  18. Analytical methods and monitoring system for E-beam flue gas treatment process

    Science.gov (United States)

    Licki, J.; Chmielewski, A. G.; Iller, E.; Zakrzewska-Trznadel, G.; Tokunaga, O.; Hashimoto, S.

    1998-06-01

    The results of reliable and precise measurement of gas composition in different key points of e-beam installation are necessary for its proper operation and control. Only the composition of flue gas coming into installation is adequate to composition of flue gas emitted from coal-fired boiler. At other points of e-b installation the gas composition is strongly modified by process conditions therefore specific measuring system (sampling and conditioning system and set of gas analyzers) for its determination are required. In the paper system for gas composition measurement at inlet and outlet of e-b installation are described. Process parameters are continuously monitoring by CEM system and occasionally by the grab sample system. Both system have been tested at pilot plant at EPS Kawȩczyn.

  19. Wearable blood flowmeter appcessory with low-power laser Doppler signal processing for daily-life healthcare monitoring.

    Science.gov (United States)

    Kuwabara, K; Higuchi, Y; Ogasawara, T; Koizumi, H; Haga, T

    2014-01-01

    A new appcessory for monitoring peripheral blood flow in daily life consists of a wearable laser Doppler sensor device and a cooperating smart phone application. Bluetooth Low Energy connects them wirelessly. The sensor device features ultralight weight of 15 g and an intermittent signal processing technique that reduces power consumption to only 7 mW at measurement intervals of 0.1 s. These features enable more than 24-h continuous monitoring of peripheral blood flow in daily life, which can provide valuable vital-sign information for healthcare services. PMID:25571431

  20. Pharmaceutical care in diabetes: quantifying and evaluating community pharmacy’s support to patients performing blood glucose self-monitoring

    OpenAIRE

    Storimans, M.J.

    2006-01-01

    The high prevalence and incidence of diabetes mellitus and its complications have led to a serious growth in the demand for diabetes-related health care in the Netherlands. Since self-management by the patient is a cornerstone of the treatment plan, self-management education has become an important part of diabetes care. One aspect of self-management education is self-monitoring of blood glucose (SMBG). SMBG is a technique that enables patients to monitor and react to changes in their blood g...

  1. Association between self-monitoring of blood glucose and diet among minority patients with diabetes*

    Science.gov (United States)

    McANDREW, Lisa M.; HOROWITZ, Carol R.; LANCASTER, Kristie J.; QUIGLEY, Karen S.; POGACH, Leonard M.; MORA, Pablo A.; LEVENTHAL, Howard

    2015-01-01

    Background Self-monitoring of blood glucose (SMBG) is used to regulate glucose control. It is unknown whether SMBG can motivate adherence to dietary recommendations. We predicted that participants who used more SMBG would also report lower fat and greater fruit and vegetable consumption. Methods The present study was a cross-sectional study of 401 primarily minority individuals living with diabetes in East Harlem, New York. Fat intake and fruit and vegetable consumption were measured with the Block Fruit/Vegetable/Fiber and Fat Screeners. Results Greater frequency of SMBG was associated with lower fat intake (rs = −0.15; P < 0.01), but not fruit and vegetable consumption. The effects of SMBG were not moderated by insulin use; thus, the relationship was significant for those individuals both on and not on insulin. A significant interaction was found between frequency of SMBG and changing one’s diet in response to SMBG on total fat intake. The data suggest that participants who use SMBG to guide their diet do not have to monitor multiple times a day to benefit. Conclusion The present study found that the frequency of SMBG was associated with lower fat intake. Patients are often taught to use SMBG to guide their self-management. This is one of the first studies to examine whether SMBG is associated with better dietary intake. PMID:21599868

  2. Remote monitoring of gas supply systems; Fernueberwachung von Gasversorgungssystemen

    Energy Technology Data Exchange (ETDEWEB)

    Neubeck, K. [WIKA Alexander Wiegand GmbH und Co., Klingenberg (Germany). Bereich Marketing; Weber, H. [Weber Engineering, Planungsbuero fuer Automatisierungstechnik, Friedberg (Germany)

    1997-11-01

    The data transmission service `Modacom` is suited for the establishment of permanent remote monitoring systems with reliable full-scale data communication. The contribution presents the example of a fuelling system, but many other applications are possible. (orig.) [Deutsch] Mit dem Datenfunkdienst `Modacom` laesst sich ein permanentes Fernueberwachungssystem aufbauen - und das mit flaechendeckender und uebertragungssicherer Datenkommunikation. Am Beispiel einer universell einsetzbaren Fernwirkunterstation mit dem dazugehoerigen Prozessleitsystem lassen sich Tankanlagen effizient managen. Doch sind der Anwendung keine Grenzen gesetzt, denn das System ist an viele Applikationen anpassbar. (orig.)

  3. Applications of optical measurement technology in pollution gas monitoring at thermal power plants

    Science.gov (United States)

    Wang, Jian; Yu, Dahai; Ye, Huajun; Yang, Jianhu; Ke, Liang; Han, Shuanglai; Gu, Haitao; Chen, Yingbin

    2011-11-01

    This paper presents the work of using advanced optical measurement techniques to implement stack gas emission monitoring and process control. A system is designed to conduct online measurement of SO2/NOX and mercury emission from stacks and slipping NH3 of de-nitrification process. The system is consisted of SO2/NOX monitoring subsystem, mercury monitoring subsystem, and NH3 monitoring subsystem. The SO2/NOX monitoring subsystem is developed based on the ultraviolet differential optical absorption spectroscopy (UV-DOAS) technique. By using this technique, a linearity error less than +/-1% F.S. is achieved, and the measurement errors resulting from optical path contamination and light fluctuation are removed. Moreover, this subsystem employs in situ extraction and hot-wet line sampling technique to significantly reduce SO2 loss due to condensation and protect gas pipeline from corrosion. The mercury monitoring subsystem is used to measure the concentration of element mercury (Hg0), oxidized mercury (Hg2+), and total gaseous mercury (HgT) in the flue gas exhaust. The measurement of Hg with a low detection limit (0.1μg/m3) and a high sensitivity is realized by using cold vapor atom fluorescence spectroscopy (CVAFS) technique. This subsystem is also equipped with an inertial separation type sampling technique to prevent gas pipeline from being clogged and to reduce speciation mercury measurement error. The NH3 monitoring subsystem is developed to measure the concentration of slipping NH3 and then to help improving the efficiency of de-nitrification. The NH3 concentration as low as 0.1ppm is able to be measured by using the off-axis integrated cavity output spectroscopy (ICOS) and the tunable diode laser absorption spectroscopy (TDLAS) techniques. The problem of trace NH3 sampling loss is solved by applying heating the gas pipelines when the measurement is running.

  4. Traditional transcutaneous approaches in head and neck surgery [

    Directory of Open Access Journals (Sweden)

    Goessler, Ulrich R.

    2012-12-01

    Full Text Available [english] The treatment of laryngeal and hypopharyngeal malignancies remains a challenging task for the head and neck surgeon as the chosen treatment modality often has to bridge the gap between oncologically sound radicality and preservation of function. Due to the increase in transoral laser surgery in early tumor stages and chemoradiation in advanced stages, the usage of traditional transcutaneous approaches has decreased over the recent past. In addition, the need for a function-sparing surgical approach as well as highest possible quality of life has become evident. In view of these facts, rationale and importance of traditional transcutaneous approaches to the treatment of laryngeal and hypopharyngeal malignancies are discussed in a contemporary background. The transcutaneous open partial laryngectomies remain a valuable tool in the surgeon's armamentarium for the treatment of early and advanced laryngeal carcinomas, especially in cases of impossible laryngeal overview using the rigid laryngoscope. Open partial laryngetomies offer superior overview and oncologic safety at the anterior commissure, especially in recurrencies. In select advanced cases and salvage settings, the supracricoid laryngectomy offers a valuable tool for function-preserving but oncologically safe surgical therapy at the cost of high postoperative morbidity and a very demanding rehabilitation of swallowing.In hypopharyngeal malignancies, the increasing use of transoral laser surgery has led to a decline in transcutaneous resections via partial pharyngectomy with partial laryngectomy in early tumor stages. In advanced stages of tumors of the piriform sinus and the postcricoid area with involvement of the larynx, total laryngectomy with partial pharyngectomy is an oncologically safe approach. The radical surgical approach using circumferent laryngopharyngectomy with/without esophagectomy is indicated in salvage cases with advanced recurrences or as a primary surgical

  5. An hour-specific transcutaneous bilirubin nomogram for Mongolian neonates

    OpenAIRE

    Akahira-Azuma, Moe; Yonemoto, Naohiro; Mori, Rintaro; Hosokawa, Shinichi; Matsushita, Takeji; Sukhbat, Khulan; Nansal, Gerelmaa; Bavuusuren, Bayasgalantai; Shonkhuuz, Enkhtur

    2015-01-01

    Transcutaneous bilirubin (TcB) nomograms have been developed for different populations. However, the TcB level, rate of rise and peak varies among countries and ethnicities. The aim of this study was to establish an hour-specific TcB nomogram for healthy term and late preterm Mongolian neonates during the first 144 h after birth. A total of 5084 TcB measurements from 1297 healthy neonates (gestational age ≥35 weeks, birth weight ≥2000 g) were obtained from October 2012 to October 2013. All me...

  6. Risk-adjusted monitoring of blood-stream infection in paediatric intensive care: a data linkage study

    OpenAIRE

    Harron, K.; WADE, A.; Muller-Pebody, B; Goldstein, H.; Parslow, R.; Gray, J.; Hartley, J. C.; Mok, Q; Gilbert, R.

    2013-01-01

    PURPOSE: National monitoring of variation in the quality of infection control in paediatric intensive care units (PICUs) requires comparisons of risk-adjusted rates. To inform the development of a national monitoring system, we evaluated the effects of risk-adjustment and outcome definition on comparisons of blood-stream infection (BSI) rates in PICU, using linkage of risk-factor data captured by national audit (PICANet) with laboratory records of BSI. METHODS: Admission data for two children...

  7. French wholesale electricity and gas markets in 2007. Monitoring report

    International Nuclear Information System (INIS)

    Settled on 24 March 2000, the French Energy Regulatory Commission (CRE) is an independent authority. CRE supports an efficient functioning of the electricity and natural gas markets, to the advantage of final consumers. CRE ensures the absence of any discrimination, cross subsidy or obstacle to competition. CRE has examined the functioning of the wholesale markets for electricity and gas for the year 2007. The report discloses the results of a first set of analyses, as well as the next actions to be conducted in order to explain some observed behaviour. On the basis of analyses led on the electricity market, CRE observes that nuclear generation was marginal during 15% of the hours in 2007. Hydraulic generation was marginal during 25% of the hours, coal generation during 25% to 30% of the hours and oil generation during less than 2% of the hours in 2007. Finally, prices on the border markets were of high importance on the French prices which they determined during 20% to 25% of the hours of the year. On an efficient market, the price is driven by the marginal power plant among all those which contribute to satisfy the demand. Therefore the wholesale price formation is conditioned by the frequency of marginality of each generation sector and by the associated valuation of production. On the day-ahead market, when nuclear or hydraulic generation was marginal, the day-ahead price reflected the associated valuation which was decided by EDF. The level of this valuation was generally higher than the marginal generation cost of these plants. As a matter of fact, a producer, even market dominant, may legitimately seek to optimise its income, provided that there is no abuse of a dominant position or any price manipulation. Consequently, at this stage, the relevance of the valuation method for nuclear and hydraulic generation on the wholesale market is still to be checked. CRE has also demonstrated that generation transparency, even if it gradually increased, still needs to

  8. A plunger lift and monitoring system for gas wells based on deployment-retrievement integration

    OpenAIRE

    Zheng Tong; Xiaohan Pei; Zejun Shen; Zhongxian Hao; Haifeng Niu

    2015-01-01

    As a necessary step, removing liquid in the wellbore plays an important role during the production of gas wells. Plunger lift is a widely-used intermittent deliquification process for gas wells. However, the manual control way and wire logging are still utilized as a downhole monitoring way for plunger lift, which is not efficient in terms of interrupting the production. This paper presents an improved solution that logging instruments canister are deployed and retrieved by means of a new ass...

  9. Hierarchical Leak Detection and Localization Method in Natural Gas Pipeline Monitoring Sensor Networks

    OpenAIRE

    Ning Yu; Renjian Feng; Jiangwen Wan; Yinfeng Wu; Yang Yu

    2011-01-01

    In light of the problems of low recognition efficiency, high false rates and poor localization accuracy in traditional pipeline security detection technology, this paper proposes a type of hierarchical leak detection and localization method for use in natural gas pipeline monitoring sensor networks. In the signal preprocessing phase, original monitoring signals are dealt with by wavelet transform technology to extract the single mode signals as well as characteristic parameters. In the initia...

  10. Effect of caffeine citrate on the blood gas indexes in newborns with apnea

    Institute of Scientific and Technical Information of China (English)

    Chun-Yan Wu; Xiao-Jiao Wang; Yuan-Yuan Wang

    2015-01-01

    Objective:To explore the effect of caffeine citrate on the blood gas indexes in newborns with apnea.Methods: A total of 108 primary premature infants diagnosed with apnea were included in the study and randomized into the caffeine citrate treatment group and the aminophylline control group with 54 cases in each group. The apnea episode duration 72 h after medication, and the blood gas before medication and 24, 48, 72 h after medication between the two groups were compared.Results: The apnea episode duration 24, 48, 72 h after medication in the treatment group was significantly shorter than that in the control group. PaO2 24, 48, 72 h after medication in the treatment group was significantly higher than that in the control group, while PaCO2 was significantly reduced when compared with the control group.Conclusions:Caffeine citrate has achieved a significantly clinical effect in the treatment of newborn apnea, as a new therapeutic drug, is worthy of being widely recommended in the clinic.

  11. Influence of transcutaneous electrical stimulation on heterotopic ossification: an experimental study in Wistar rats

    Directory of Open Access Journals (Sweden)

    T.G.G. Zotz

    2015-01-01

    Full Text Available Heterotopic ossification (HO is a metaplastic biological process in which there is newly formed bone in soft tissues, resulting in joint mobility deficit and pain. Different treatment modalities have been tried to prevent HO development, but there is no consensus on a therapeutic approach. Since electrical stimulation is a widely used resource in physiotherapy practice to stimulate joint mobility, with analgesic and anti-inflammatory effects, its usefulness for HO treatment was investigated. We aimed to identify the influence of electrical stimulation on induced HO in Wistar rats. Thirty-six male rats (350-390 g were used, and all animals were anesthetized for blood sampling before HO induction, to quantify the serum alkaline phosphatase. HO induction was performed by bone marrow implantation in both quadriceps of the animals, which were then divided into 3 groups: control (CG, transcutaneous electrical nerve stimulation (TENS group (TG, and functional electrical stimulation (FES group (FG with 12 rats each. All animals were anesthetized and electrically stimulated twice per week, for 35 days from induction day. After this period, another blood sample was collected and quadriceps muscles were bilaterally removed for histological and calcium analysis and the rats were killed. Calcium levels in muscles showed significantly lower results when comparing TG and FG (P

  12. Methodology and technology for peripheral and central blood pressure and blood pressure variability measurement: current status and future directions - Position statement of the European Society of Hypertension Working Group on blood pressure monitoring and cardiovascular variability.

    Science.gov (United States)

    Stergiou, George S; Parati, Gianfranco; Vlachopoulos, Charalambos; Achimastos, Apostolos; Andreadis, Emanouel; Asmar, Roland; Avolio, Alberto; Benetos, Athanase; Bilo, Grzegorz; Boubouchairopoulou, Nadia; Boutouyrie, Pierre; Castiglioni, Paolo; de la Sierra, Alejandro; Dolan, Eamon; Head, Geoffrey; Imai, Yutaka; Kario, Kazuomi; Kollias, Anastasios; Kotsis, Vasilis; Manios, Efstathios; McManus, Richard; Mengden, Thomas; Mihailidou, Anastasia; Myers, Martin; Niiranen, Teemu; Ochoa, Juan Eugenio; Ohkubo, Takayoshi; Omboni, Stefano; Padfield, Paul; Palatini, Paolo; Papaioannou, Theodore; Protogerou, Athanasios; Redon, Josep; Verdecchia, Paolo; Wang, Jiguang; Zanchetti, Alberto; Mancia, Giuseppe; O'Brien, Eoin

    2016-09-01

    Office blood pressure measurement has been the basis for hypertension evaluation for almost a century. However, the evaluation of blood pressure out of the office using ambulatory or self-home monitoring is now strongly recommended for the accurate diagnosis in many, if not all, cases with suspected hypertension. Moreover, there is evidence that the variability of blood pressure might offer prognostic information that is independent of the average blood pressure level. Recently, advancement in technology has provided noninvasive evaluation of central (aortic) blood pressure, which might have attributes that are additive to the conventional brachial blood pressure measurement. This position statement, developed by international experts, deals with key research and practical issues in regard to peripheral blood pressure measurement (office, home, and ambulatory), blood pressure variability, and central blood pressure measurement. The objective is to present current achievements, identify gaps in knowledge and issues concerning clinical application, and present relevant research questions and directions to investigators and manufacturers for future research and development (primary goal). PMID:27214089

  13. Blood gas analysis of patients with malignant ascites in 42 cases%恶性腹水患者42例血气分析

    Institute of Scientific and Technical Information of China (English)

    谭斌; 辛瑞娟; 陈慧; 冯雅宁; 白飞虎

    2015-01-01

    Objective:To explore the clinical significance of blood gas analysis indexes of patients with malignant ascites.Methods:The blood gas parameters of the malignant ascites group and the healthy control group were detected by using automatic blood gas analyzer.The blood sodium(Na+),the blood chlorine(Cl-) concentration were detected in the venous blood.Results:pH value,PCO2, PO2 and HCO3- of the malignant ascites group were compared with those of the control group,and the differences were all statistically significant(P<0.05).The incidence rate of hypoxemia in the malignant ascites group was 69.0%,the incidence rate of acid-base imbalance was 66.7%,the acid-base imbalance was give priority to with simple respiratory alkalosis,the incidence rate was 44.89%.Conclusion:Patients with malignant ascites easily have acid-base imbalance and hypoxia,the acid-base imbalance is give priority to with respiratory alkalosis.It should monitor the blood gas analysis in a timely manner and give timely symptomatic treatment to improve the life quality of the patients.%目的:探讨恶性腹水患者血气分析指标的临床意义。方法:采用全自动血气分析仪检测恶性腹水组和健康对照组血气参数,静脉血测定血钠(Na+)、血氯(Cl-)浓度。结果:恶性腹水组pH值、PCO2、PO2和HCO3-与对照组比较,差异均有统计学意义。恶性腹水组低氧血症的发生率69.0%,酸碱失衡的发生率66.7%,酸碱失衡以单纯性呼吸性碱中毒为主,发生率44.89%。结论:恶性腹水患者易发生酸碱失衡及低氧血症,酸碱失衡以呼吸性碱中毒为主,应及时进行血气分析,及时对症治疗,提高患者生活质量。

  14. Systematic study of RPC performances in polluted or varying gas mixtures compositions: an online monitor system for the RPC gas mixture at LHC

    CERN Document Server

    Capeans, M; Mandelli, B

    2012-01-01

    The importance of the correct gas mixture for the Resistive Plate Chamber (RPC) detector systems is fundamental for their correct and safe operation. A small change in the percentages of the gas mixture components can alter the RPC performance and this will rebound on the data quality in the ALICE, ATLAS and CMS experiments at CERN. A constant monitoring of the gas mixture injected in the RPCs would avoid such kind of problems. A systematic study has been performed to understand RPC performances with several gas mixture compositions and in the presence of common gas impurities. The systematic analysis of several RPC performance parameters in different gas mixtures allows the rapid identification of any variation in the RPC gas mixture. A set-up for the online monitoring of the RPC gas mixture in the LHC gas systems is also proposed.

  15. Design and Initial Tests of a Gas Scattering Energy Monitor in the PEFP RFQ and DTL

    CERN Document Server

    Han, Sang-Hyo

    2005-01-01

    We have developed a gas scattering energy monitor to measure the energy spectrum of the proton beam at the end of 3MeV RFQ and 20MeV DTL for the low energy part of the PEFP(Proton Engineering Frontier Project). The energy monitor is comprised of a Xe scattering chamber, two collimators to reduce the beam intensity, and a surface barrier detector for measuring the proton energy. In order to measure the beam current simultaneously, a faraday cup is incorporated into the energy monitor. The calculated flux attenuation through the 0.2 mm diameter collimator is 3·10-4

  16. In Vivo Monitoring of Hemodynamic Changes during Clogging and Unclogging of Blood Supply for the Application of Clinical Shock Detection

    Science.gov (United States)

    Kanawade, Rajesh; Stelzle, Florian; Schmidt, Michael

    This paper presents a novel methodology in early detection of clinical shock by monitoring hemodynamic changes using diffuse reflectance measurement technique. Detailed prototype of the reflectance measurement system and data analysis technique of hemodynamic monitoring was carried out in our laboratory. The real time in-vivo measurements were done from the index finger. This study demonstrates preliminary results of real time monitoring of reduced/- oxyhemoglobin changes during clogging and unclogging of blood flow in the finger tip. The obtained results were verified with pulse-oximeter values, connected to the tip of the same index finger.

  17. Role of therapeutic drug monitoring in pulmonary infections : use and potential for expanded use of dried blood spot samples

    NARCIS (Netherlands)

    Hofman, Susan; Bolhuis, Mathieu S.; Koster, Remco A.; Akkerman, Onno W.; van Assen, Sander; Stove, Christophe; Alffenaar, Jan-Willem C.

    2015-01-01

    Respiratory tract infections are among the most common infections in men. We reviewed literature to document their pharmacological treatments, and the extent to which therapeutic drug monitoring (TDM) is needed during treatment. We subsequently examined potential use of dried blood spots as sample p

  18. Implementation of Out-of-Office Blood Pressure Monitoring in the Netherlands: From Clinical Guidelines to Patients’ Adoption of Innovation

    NARCIS (Netherlands)

    Carrera, Pricivel M.; Lambooij, Mattijs S.

    2015-01-01

    Out-of-office blood pressure monitoring is promoted by various clinical guidelines toward properly diagnosing and effectively managing hypertension and engaging the patient in their care process. In the Netherlands, however, the Dutch cardiovascular risk management (CVRM) guidelines do not explicitl

  19. Performance in real condition of photonic crystal sensor based NO2 gas monitoring system

    Science.gov (United States)

    Rahmat, M.; Maulina, W.; Rustami, E.; Azis, M.; Budiarti, D. R.; Seminar, K. B.; Yuwono, A. S.; Alatas, H.

    2013-11-01

    In this report we discuss the performance in real condition of an optical based real-time NO2 gas monitoring system. For detecting the gas concentration in the ambient air we have developed an optical sensor based on one-dimensional photonic crystal with two defects that allows the existence of photonic pass band inside the associated photonic band gap. To measure the gas concentration, we dissolve the corresponding NO2 gas into a specific Griess Saltzman reagent solution. The change of gas concentration in the related dissolved-solution can be inspected by the photonic pass band peak variation. It is observed that the wavelength of the photonic pass band peak of the fabricated photonic crystal is nearly coincide with the wavelength of the associated solution highest absorbance. The laboratory test shows that the device works properly, whereas the field measurement test demonstrates accurate results with validation error of 1.56%.

  20. A highlight on lipid based nanocarriers for transcutaneous immunization.

    Science.gov (United States)

    Nasr, Maha; Abdel-Hamid, Sameh; Alyoussef, Abdullah A

    2015-01-01

    Transcutaneous vaccination has become a widely used technique for providing immunity against several types of pathogens, taking advantage of the immune components found in the skin. The success in the field of vaccination has not only relied on the type of antigen and adjuvant delivered, but also on how they are delivered. In this regard, particulate carriers, especially nanoparticles have evoked considerable interest, owing to the desirable properties that they impart to the substance being delivered. The presentation of antigens by the nanoparticles mimics the presentation of the immunogen by the pathogen; hence, it creates a similar immune response. Furthermore, nanoparticles protect the antigen from degradation and allow its prolonged release, which maximizes its exposure to the immune cells. The most commonly used materials for the formulation of nanoparticles are either polymer-based or lipid based. This review will focus on the lipid based nanocarriers, either vesicular such as liposomes, transfersomes, and ethosomes, or non-vesicular such as cubosomes, solid lipid nanoparticles, nano-structured lipid carriers, solid in oil nanodispersions, lipoplexes, and hybrid polymeric-lipidic systems. The applications of these carriers in the field of transcutaneous immunization will be discussed in this review as well. PMID:25658381

  1. Using natural, stable calcium isotopes of human blood to detect and monitor changes in bone mineral balance.

    Science.gov (United States)

    Channon, Melanie B; Gordon, Gwyneth W; Morgan, Jennifer L L; Skulan, Joseph L; Smith, Scott M; Anbar, Ariel D

    2015-08-01

    We are exploring variations in the Ca isotope composition of blood and urine as a new tool for early diagnosis and monitoring of changes in bone mineral balance for patients suffering from metabolic bone disease, cancers that originate in or metastasize to bone, and for astronauts who spend time in low gravity environments. Blood samples are often collected instead of, or in addition to, urine in clinical settings, so it is useful to know if variations in the Ca isotope composition of blood carry the same information as variations in urine. We found that the Ca isotope composition of blood shifts in the same direction and to the same magnitude (~2 parts per ten thousand--pptt) as that of urine in response to skeletal unloading during bed rest. However, the Ca isotope composition of blood is lighter than that of urine by 12 ± 2 pptt. This offset between blood and urine may result from Ca isotope fractionation occurring in the kidneys. This is the first study to confirm the suspected offset between the Ca isotope composition of blood and urine in humans, to directly quantify its magnitude, and to establish that either blood or urine can be used to detect and quantify bone loss.

  2. Blood Glucose Control Using a Novel Continuous Blood Glucose Monitor and Repetitive Intravenous Insulin Boluses: Exploiting Natural Insulin Pulsatility as a Principle for a Future Artificial Pancreas

    Directory of Open Access Journals (Sweden)

    Nils K. Skjaervold

    2013-01-01

    Full Text Available The aim of this study was to construct a glucose regulatory algorithm by employing the natural pulsatile pattern of insulin secretion and the oscillatory pattern of resting blood glucose levels and further to regulate the blood glucose level in diabetic pigs by this method. We developed a control algorithm based on repetitive intravenous bolus injections of insulin and combined this with an intravascular blood glucose monitor. Four anesthetized pigs were used in the study. The animals developed a mildly diabetic state from streptozotocin pretreatment. They were steadily brought within the blood glucose target range of 4.5–6.0 mmol/L in 21 to 121 min and kept within that range for 128 to 238 min (hypoglycemic values varied from 2.9 to 51.1 min. The study confirmed our hypotheses regarding the feasibility of this new principle for blood glucose control, and the algorithm was constantly improved during the study to produce the best results in the last animals. The main obstacles were the drift of the IvS-1 sensor and problems with the calibration procedure, which calls for an improvement in the sensor stability before this method can be applied fully in new studies in animals and humans.

  3. Design layout for gas monitoring system II (GMS-2) computer system

    Energy Technology Data Exchange (ETDEWEB)

    Vo, V.; Philipp, B.L.; Manke, M.P.

    1995-08-02

    This document provides a general overview of the computer systems software that perform the data acquisition and control for the 241-SY-101 Gas Monitoring System II (GMS-2). It outlines the system layout, and contains descriptions of components and the functions they perform. The GMS-2 system was designed and implemented by Los Alamos National Laboratory and supplied to Westinghouse Hanford Company

  4. Design layout for gas monitoring system II, GMS-2, computer system

    International Nuclear Information System (INIS)

    This document provides a general overview of the computer systems software that perform the data acquisition and control for the 241-SY-101 Gas Monitoring System II (GMS-2). It outlines the system layout, and contains descriptions of components and the functions they perform. The GMS-2 system was designed and implemented by Los Alamos National Laboratory and supplied to Westinghouse Hanford Company

  5. A BEAM PROFILE MONITOR USING THE IONIZATION OF RESIDUAL-GAS IN THE BEAM PIPE

    NARCIS (Netherlands)

    SCHIPPERS, JM; KIEWIET, HH; ZIJLSTA, J

    1991-01-01

    A beam profile monitor for high energy beams, which has no intercepting parts in the beam pipe, is described. It makes use of the ionization of the residual gas, which is still present in the vacuum chamber of the beam guiding system. The detection of the ionization products is performed with microc

  6. Feasibility of monitoring gas hydrate production with time-lapse VSP

    Energy Technology Data Exchange (ETDEWEB)

    Kowalsky, M.B.; Nakagawa, S.; Moridis, G.J.

    2009-11-01

    In this work we begin to examine the feasibility of using time-lapse seismic methods-specifically the vertical seismic profiling (VSP) method-for monitoring changes in hydrate accumulations that are predicted to occur during production of natural gas.

  7. Control and monitoring of landfill gas underground migration at the City of Montreal sanitary landfill site

    International Nuclear Information System (INIS)

    The proposed paper covers the various aspects of control and monitoring of potential landfill gas (LFG) migration through soil voids or rock fractures at the City of Montreal sanitary landfill site. It depicts the social, geographical and geological context and presents a brief history of the landfill site. It describes the LFG collecting system and LFG migration monitoring equipment and programs. Finally it presents monitoring data taken over last few years. The landfill site is located in a well populated urban area. Since 1968, about 33 million metric tons of domestic and commercial waste have been buried in a former limestone quarry. Because of houses and buildings in the vicinity, 100 m in some locations, LFG underground migration is a major risk. LFG could indeed infiltrate buildings and reach explosive concentrations. So it must be controlled. The City of Montreal acquired the site in 1988 and has progressively built a LFG collecting system, composed of more than 288 vertical wells, to pump out of the landfill 280 million m3 of gas annually. To verify the efficiency of this system to minimize LFG underground migration, monitoring equipment and programs have also been designed and put into operation. The monitoring network, located all around the landfill area, is composed of 21 well nests automated to monitor presence of gas in the ground in real time. In addition, 55 individual wells, where manual measurements are made, are also available. To complete the monitoring program, some measurements are also taken in buildings, houses and underground utilities in the neighborhood of the site. Monitoring data show that LFG underground migration is well controlled. They also indicate significant decrease of migration over the years corresponding to improvements to the LFG collecting system

  8. A System for Simple Real-Time Anastomotic Failure Detection and Wireless Blood Flow Monitoring in the Lower Limbs

    Science.gov (United States)

    Rothfuss, Michael A.; Franconi, Nicholas G.; Unadkat, Jignesh V.; Gimbel, Michael L.; STAR, Alexander; Mickle, Marlin H.

    2016-01-01

    Current totally implantable wireless blood flow monitors are large and cannot operate alongside nearby monitors. To alleviate the problems with the current monitors, we developed a system to monitor blood flow wirelessly, with a simple and easily interpretable real-time output. To the best of our knowledge, the implanted electronics are the smallest in reported literature, which reduces bio-burden. Calibration was performed across realistic physiological flow ranges using a syringe pump. The device’s sensors connected directly to the bilateral femoral veins of swine. For each 1 min, blood flow was monitored, then, an occlusion was introduced, and then, the occlusion was removed to resume flow. Each vein of four pigs was monitored four times, totaling 32 data collections. The implant measured 1.70 cm3 without battery/encapsulation. Across its calibrated range, including equipment tolerances, the relative error is less than ±5% above 8 mL/min and between −0.8% and +1.2% at its largest calibrated flow rate, which to the best of our knowledge is the lowest reported in the literature across the measured calibration range. The average standard deviation of the flow waveform amplitude was three times greater than that of no-flow. Establishing the relative amplitude for the flow and no-flow waveforms was found necessary, particularly for noise modulated Doppler signals. Its size and accuracy, compared with other microcontroller-equipped totally implantable monitors, make it a good candidate for future tether-free free flap monitoring studies. PMID:27730016

  9. New Hadron Monitor By Using A Gas-Filled RF Resonator

    Energy Technology Data Exchange (ETDEWEB)

    Yonehara, Katsuya [Fermilab; Fasce, Giorgio [ECONA, Rome; Flanagan, Gene [MUONS Inc., Batavia; Johnson, Rolland [MUONS Inc., Batavia; Tollestrup, Alvin [Fermilab; Zwaska, Robert [Fermilab

    2015-05-01

    It is trend to build an intense neutrino beam facility for the fundamental physics research, e.g. LBNF at Fermilab, T2K at KEK, and CNGS at CERN. They have investigated a hadron monitor to diagnose the primary/secondary beam quality. The existing hadron monitor based on an ionization chamber is not robust in the high-radiation environment vicinity of MW-class secondary particle production targets. We propose a gas-filled RF resonator to use as the hadron monitor since it is simple and hence radiation robust in this environment. When charged particles pass through the resonator they produce ionized plasma via the Coulomb interaction with the inert gas. The beam-induced plasma changes the permittivity of inert gas. As a result, a resonant frequency in the resonator shifts with the amount of ionized electrons. The radiation sensitivity is adjustable by the inert gas pressure and the RF amplitude. The hadron profile will be reconstructed with a tomography technique in the hodoscope which consists of X, Y, and theta layers by using a strip-shaped gas resonator. The sensitivity and possible system design will be shown in this presentation.

  10. The association of serum gamma-glutamyl transpeptidase level and other laboratory parameters with blood pressure in hypertensive patients under ambulatory blood pressure monitoring

    Science.gov (United States)

    Ortakoyluoglu, Adile; Boz, Betul; Dizdar, Oguzhan Sıtkı; Avcı, Deniz; Cetinkaya, Ali; Baspınar, Osman

    2016-01-01

    Background Hypertension is a very important cause of morbidity and mortality. Serum gamma-glutamyl transpeptidase (GGT) is a biomarker of oxidative stress and associated with increased risk of hypertension and diabetes. The aim of this study was to evaluate the association of serum GGT level, which is an early marker of inflammation and endothelial dysfunction, with the deterioration of the diurnal rhythm of the blood pressure. Methods A total of 171 patients with hypertension were included in this study. Patients whose nighttime mean blood pressure, measured via ambulatory blood pressure monitoring, decreased between 10% and 20% compared with the daytime mean blood pressure were defined as “dippers”, whereas patients with a nighttime blood pressure decrease lower than 10% were defined as “non-dippers”. Results A total of 99 hypertensive patients (65 females/34 males) were classified as dippers and 72 patients (48 females/24 males) as non-dippers. The mean age of the non-dipper group was significantly greater than the dipper group. Serum GGT, C-reactive protein and uric acid levels were significantly higher among patients in the non-dipper group. Negative correlations were detected between GGT levels and diurnal systolic and diastolic blood pressure decreases. Conclusion Our findings revealed that GGT level was higher in the non-dipper group, and was negatively correlated with the nighttime decrease of diurnal blood pressure. C-reactive protein and uric acid levels were also higher in the non-dipper group. However, future randomized controlled prospective studies with larger patient populations are necessary to confirm our findings. PMID:27660457

  11. Designing optimal greenhouse gas monitoring networks for Australia

    Science.gov (United States)

    Ziehn, T.; Law, R. M.; Rayner, P. J.; Roff, G.

    2016-01-01

    Atmospheric transport inversion is commonly used to infer greenhouse gas (GHG) flux estimates from concentration measurements. The optimal location of ground-based observing stations that supply these measurements can be determined by network design. Here, we use a Lagrangian particle dispersion model (LPDM) in reverse mode together with a Bayesian inverse modelling framework to derive optimal GHG observing networks for Australia. This extends the network design for carbon dioxide (CO2) performed by Ziehn et al. (2014) to also minimise the uncertainty on the flux estimates for methane (CH4) and nitrous oxide (N2O), both individually and in a combined network using multiple objectives. Optimal networks are generated by adding up to five new stations to the base network, which is defined as two existing stations, Cape Grim and Gunn Point, in southern and northern Australia respectively. The individual networks for CO2, CH4 and N2O and the combined observing network show large similarities because the flux uncertainties for each GHG are dominated by regions of biologically productive land. There is little penalty, in terms of flux uncertainty reduction, for the combined network compared to individually designed networks. The location of the stations in the combined network is sensitive to variations in the assumed data uncertainty across locations. A simple assessment of economic costs has been included in our network design approach, considering both establishment and maintenance costs. Our results suggest that, while site logistics change the optimal network, there is only a small impact on the flux uncertainty reductions achieved with increasing network size.

  12. Ocean observatory networks monitor gas hydrates systems - Updates from Cascadia

    Science.gov (United States)

    Scherwath, M.; Kelley, D. S.; Moran, K.; Philip, B. T.; Roemer, M.; Riedel, M.; Solomon, E. A.; Spence, G.; Heesemann, M.

    2015-12-01

    Seafloor observatories have been installed at the Cascadia margin with a long-term (>20 year) lifespan. These observatories consist of a variety of node locations cabled back to shore for continuous power and communication to instruments via high bandwidth internet access. Ocean Networks Canada (ONC) maintains two hydrate sites at Barkley Canyon and Clayoquot Slope off Vancouver Island, and the Ocean Observatories Initiative (OOI) Cabled Array connects to Hydrate Ridge off the Oregon coast. Together, these installations comprise a diverse suite of different experiments. For example, a seafloor crawler, operated by Jacobs University in Bremen, travels around the Barkley hydrate mounds on a daily basis and carries out a suite of measurements such as determining the rate of change of the benthic community composition. Another example is from several years of hourly sonar data showing gas bubbles rising from the seafloor near the Bullseye Vent with varying intensities, allowing statistically sound correlations with other seafloor parameters such as ground shaking, temperature and pressure variations and currents, where tidal pressure appearing as the main driver. The Southern Hydrate Ridge is now equipped with the world's first long-term seafloor mass spectrometer, co-located with a camera and lights, hydrophone, current meters, pressure sensor, autonomous dissolved oxygen and fluid samplers, and is surrounded by a seismometer array for local seismicity. In the future, long-term data will be continuously captured and made available throughout the year covering the full range of variations of the dynamic hydrate system, and expect additional experiments to be connected to the observatories from the broader research community.

  13. Designing optimal greenhouse gas monitoring networks for Australia

    Directory of Open Access Journals (Sweden)

    T. Ziehn

    2015-08-01

    Full Text Available Atmospheric transport inversion is commonly used to infer greenhouse gas (GHG flux estimates from concentration measurements. The optimal location of ground based observing stations that supply these measurements can be determined by network design. Here, we use a Lagrangian particle dispersion model (LPDM in reverse mode together with a Bayesian inverse modelling framework to derive optimal GHG observing networks for Australia. This extends the network design for carbon dioxide (CO2 performed by Ziehn et al. (2014 to also minimize the uncertainty on the flux estimates for methane (CH4 and nitrous oxide (N2O, both individually and in a combined network using multiple objectives. Optimal networks are generated by adding up to 5 new stations to the base network, which is defined as two existing stations, Cape Grim and Gunn Point, in southern and northern Australia respectively. The individual networks for CO2, CH4 and N2O and the combined observing network show large similarities because the flux uncertainties for each GHG are dominated by regions of biologically productive land. There is little penalty, in terms of flux uncertainty reduction, for the combined network compared to individually designed networks. The location of the stations in the combined network is sensitive to variations in the assumed data uncertainty across locations. A simple assessment of economic costs has been included in our network design approach, considering both establishment and maintenance costs. Our results suggest that while site logistics change the optimal network, there is only a small impact on the flux uncertainty reductions achieved with increasing network size.

  14. Satellite Monitoring Systems for Shipping and Offshore Oil and Gas Industry in the Baltic Sea

    Directory of Open Access Journals (Sweden)

    Kostianoy A.G.

    2015-06-01

    Full Text Available Shipping activities, oil production and transport in the sea, oil handled in harbors, construction and exploitation of offshore oil and gas pipelines have a number of negative impacts on the marine environment and coastal zone of the seas. In 2004-2014 we elaborated several operational satellite monitoring systems for oil and gas companies in Russia and performed integrated satellite monitoring of the ecological state of coastal waters in the Baltic, Black, Caspian, and Kara seas, which included observation of oil pollution, suspended matter, and algae bloom at a fully operational mode. These monitoring systems differ from the existing ones by the analysis of a wide spectrum of satellite, meteorological and oceanographic data, as well as by a numerical modeling of oil spill transformation and transport in real weather conditions. Our experience in the Baltic Sea includes: (1 integrated satellite monitoring of oil production at the LUKOIL-KMN Ltd. D-6 oil rig in the Southeastern Baltic Sea (Kravtsovskoe oil field in 2004-2014; (2 integrated satellite monitoring of the “Nord Stream” underwater gas pipeline construction and exploitation in the Gulf of Finland (2010-2013; (3 numerical modeling of risks of oil pollution caused by shipping along the main maritime shipping routes in the Gulf of Finland, the Baltic Proper, and in the Southeastern Baltic Sea; (4 numerical modeling of risks of oil pollution caused by oil production at D-6 oil rig and oil transportation on shore via the connecting underwater oil pipeline.

  15. Gas Filled RF Resonator Hadron Beam Monitor for Intense Neutrino Beam Experiments

    Energy Technology Data Exchange (ETDEWEB)

    Yonehara, Katsuya [Fermilab; Abrams, Robert [MUONS Inc., Batavia; Dinkel, Holly [U. Missouri, Columbia; Freemire, Ben [IIT, Chicago; Johnson, Rolland [MUONS Inc., Batavia; Kazakevich, Grigory [MUONS Inc., Batavia; Tollestrup, Alvin [Fermilab; Zwaska, Robert [Fermilab

    2016-06-01

    MW-class beam facilities are being considered all over the world to produce an intense neutrino beam for fundamental particle physics experiments. A radiation-robust beam monitor system is required to diagnose the primary and secondary beam qualities in high-radiation environments. We have proposed a novel gas-filled RF-resonator hadron beam monitor in which charged particles passing through the resonator produce ionized plasma that changes the permittivity of the gas. The sensitivity of the monitor has been evaluated in numerical simulation. A signal manipulation algorithm has been designed. A prototype system will be constructed and tested by using a proton beam at the MuCool Test Area at Fermilab.

  16. Fluorescent blood glucose monitor by hemin-functionalized graphene quantum dots based sensing system

    Energy Technology Data Exchange (ETDEWEB)

    He, Yuezhen; Wang, Xiaoxun; Sun, Jian; Jiao, Shoufeng; Chen, Hongqi; Gao, Feng; Wang, Lun, E-mail: wanglun@mail.ahnu.edu.cn

    2014-01-31

    Graphical abstract: -- Highlights: •Hemin is assembled onto the surfaces of graphene quantum dots (GQDs). •With the aid of hemin, H{sub 2}O{sub 2} could quench the FL signal of GQDs obviously. •Based on this effect, a fluorescent platform is proposed for the sensing of glucose. •The proposed method provides a new pathway to explore practical application of GQDs. -- Abstract: In the present work, a highly sensitive and specific fluorescent biosensor for blood glucose monitoring is developed based on hemin-functionalized graphene quantum dots (GQDs) and glucose oxidase (GOx) system. The GQDs which are simply prepared by pyrolyzing citric acid exhibit strong fluorescence and good water-solubility. Due to the noncovalent assembly between hemin and GQDs, the addition of hemin can make hydrogen peroxide (H{sub 2}O{sub 2}) to destroy the passivated surface of GQDs, leading to significant fluorescence quenching of GQDs. Based on this effect, a novel fluorescent platform is proposed for the sensing of glucose. Under the optimized conditions, the linear range of glucose is from 9 to 300 μM, and the limit of detection is 0.1 μM. As unique properties of GQDs, the proposed biosensor is green, simple, cost-efficient, and it is successfully applied to the determination of glucose in human serum. In addition, the proposed method provides a new pathway to further design the biosensors based on the assembly of GQDs with hemin for detection of biomolecules.

  17. Effect of breath-hold on blood gas analysis in captive Pacific white-sided dolphins (Lagenorhynchus obliquidens).

    Science.gov (United States)

    Terasawa, Fumio; Ohizumi, Hiroshi; Ohshita, Isao

    2010-09-01

    The effect of a breath-hold on blood gas was evaluated in captive Pacific white-sided dolphins (Lagenorhynchus obliquidens). Serial blood collections were performed from a vessel on the ventral surface of the flukes during breath-hold. In total, 178 blood samples were taken from three dolphins for five trials in each animal. During a breath-hold, partial pressure of oxygen (Po₂) decreased from 152.5 to 21.8 mmHg and partial pressure of carbon dioxide (Po₂) conversely increased from 31.8 to 83.6 mmHg. The range of pH was 7.54 to 7.25, suggesting drastic change from alkalemia to acidemia. These wide ranges of blood gas imply a considerable change of oxygen affinity caused by the Bohr effect during breath-hold, which enable effective uptake and distribution of oxygen to metabolizing tissues.

  18. Exhaust gas monitoring based on absorption spectroscopy in the process industry

    Science.gov (United States)

    Zhang, Shuai; Liu, Wen-qing; Zhang, Yu-jun; Shu, Xiao-wen; Kan, Rui-feng; Cui, Yi-ben; He, Ying; Xu, Zhen-yu; Geng, Hui; Liu, Jian-guo

    2009-07-01

    This non-invasive gas monitor for exhaust gas monitoring must has high reliability and requires little maintenance. Monitor for in-situ measurements using tunable diode laser absorption spectroscopy (TDLAS) in the near infrared, can meet these requirements. TDLAS has evolved over the past decade from a laboratory especially to an accepted, robust and reliable technology for trace gas sensing. With the features of tunability and narrow linewidth of the distributed feedback (DFB) diode laser and by precisely tuning the laser output wavelength to a single isolated absorption line of the gas, TDLAS technique can be utilized to measure gas concentration with high sensitivity. Typical applications for monitoring of H2S, NH3, HC1 and HF are described here together by wavelength modulation spectroscopy with second-harmonic(WMS-2F) detection. This paper will illustrate the problems related to on-line applications, in particular, the overfall effects, automatic light intensity correction, temperature correction, which impacted on absorption coefficient and give details of how effect of automatic correction is necessary. The system mainly includes optics and electronics, optical system mainly composed of fiber, fiber coupler and beam expander, the electron part has been placed in safe analysis room not together with the optical part. Laser merely passes through one-meter-long pipes by the fiber coupling technology, so the system itself has anti-explosion. The results of the system are also presented in the end, the system's response time is only 0.5s, and can be achieved below 1×10-5 the detection limit at the volume fraction, it can entirely replace the traditional methods of detection exhaust gas in the process industry.

  19. Artificial neural networks for monitoring the gas turbine; Artificiella neuronnaet foer gasturbinoevervakning

    Energy Technology Data Exchange (ETDEWEB)

    Fast, Magnus; Thern, Marcus [Inst. foer Energivetenskaper, Lunds Univ. (Sweden)

    2011-10-15

    Through available historical operational data from gas turbines, fast, accurate, easy to use and reliable models can be developed. These models can be used for monitoring of gas turbines and assist in the transition from today's time-based maintenance to condition based maintenance. For the end user this means that, because only operational data is needed, they can easily develop their own tools independent of the manufacturer. Traditionally these types of models are constructed with physical relations for e.g., mass, energy and momentum. To develop a model with physical relations is often laborious and requires classified information which the end user does not have access to. Research has shown that by producing models using operational data a very high model precision can be achieved. When implementing these models in a power plant computer system the gas turbine's performance can be monitored in real time. This can facilitate fault detection at an early stage, and if necessary, stop the gas turbine before major damage occurs. For the power plant owner, this means that the gas turbine reliability is increased since the need for maintenance is minimized and the downtime is reduced. It also means that a measure of the gas turbine's overall status is continuously available, with respect to e.g. degradation, which helps in the planning of service intervals. The tool used is called artificial neural networks (ANN), a collective name for a number of algorithms for information processing that attempts to mimic the nerve cell function. Just like real networks of neurons in a brain, these artificial neural networks have the ability to learn. In this case, neural networks are trained to mimic the behavior of gas turbines by introducing them to data from real gas turbines. After a neural network is trained it represents a very accurate model of the gas turbine that it is trained to emulate.

  20. On-line monitoring of dissolved gas-in-oil with FTIR spectra

    Institute of Scientific and Technical Information of China (English)

    Xianyong Liu; Yunluo Liu; Li Yue

    2003-01-01

    To overcome the disadvantages of conventional DGA (dissolved gas-in-oil) analysis using gas chromatography and other electrochemical sensors, initial researches were completed to realize on-line monitoring of dissolved gas-in-oil of power transformers using FTIR (Fourier Transform InfraRed) spectroscopy. Gas cell method is used to determine the characteristic absorption peaks of each diagnostic gas; simple and novel devices and procedures were designed in order to get measurable samples and spectra of mixed diagnostic gases with known concentration are taken using long optical path gas cell. The range of wavelength is estimated to be 3.0-13.9 μm from experimental spectra data. Hence the corresponding sampling frequency range should be in 536-4288 Hz and usable optical materials are suggested. It is concluded that a resolution of 10 cm-1 may well satisfy the monitoring of all diagnostic gases and water content except hydrogen, and the lowest detection limit may be as low as 2×l0-8 to acetylene with a 2.4-meter-long optical length.

  1. Enabling Technology for Monitoring & Predicting Gas Turbine Health & Performance in IGCC Powerplants

    Energy Technology Data Exchange (ETDEWEB)

    Kenneth A. Yackly

    2005-12-01

    The ''Enabling & Information Technology To Increase RAM for Advanced Powerplants'' program, by DOE request, was re-directed, de-scoped to two tasks, shortened to a 2-year period of performance, and refocused to develop, validate and accelerate the commercial use of enabling materials technologies and sensors for coal/IGCC powerplants. The new program was re-titled ''Enabling Technology for Monitoring & Predicting Gas Turbine Health & Performance in IGCC Powerplants''. This final report summarizes the work accomplished from March 1, 2003 to March 31, 2004 on the four original tasks, and the work accomplished from April 1, 2004 to July 30, 2005 on the two re-directed tasks. The program Tasks are summarized below: Task 1--IGCC Environmental Impact on high Temperature Materials: The first task was refocused to address IGCC environmental impacts on high temperature materials used in gas turbines. This task screened material performance and quantified the effects of high temperature erosion and corrosion of hot gas path materials in coal/IGCC applications. The materials of interest included those in current service as well as advanced, high-performance alloys and coatings. Task 2--Material In-Service Health Monitoring: The second task was reduced in scope to demonstrate new technologies to determine the inservice health of advanced technology coal/IGCC powerplants. The task focused on two critical sensing needs for advanced coal/IGCC gas turbines: (1) Fuel Quality Sensor to rapidly determine the fuel heating value for more precise control of the gas turbine, and detection of fuel impurities that could lead to rapid component degradation. (2) Infra-Red Pyrometer to continuously measure the temperature of gas turbine buckets, nozzles, and combustor hardware. Task 3--Advanced Methods for Combustion Monitoring and Control: The third task was originally to develop and validate advanced monitoring and control methods for coal/IGCC gas

  2. Monitoring of milk production and total cholesterol concentration, gamma-glutamyltransferase, and glutathione peroxidase in Simmental cows blood

    Directory of Open Access Journals (Sweden)

    Terezija Silvija Marenjak

    2007-06-01

    Full Text Available Milk production, total blood cholesterol concentration and activity ofgamma-glutamyl-transferase (GGT and glutation peroxidase (GPx in blood of Simmental cows was monitored during two control periods (first control period - May 2004; second control period - June 2004. The objective was to determine a relationship between milk production and particular blood parameters, and their possible implementation as an indicator of milk production. Twelve Simmental cows from small-scale dairy farm in Zagreb County were included in the study. Cows originated from the same herd andwere roughly of the same parity and stage of lactation. The average milk production and total blood cholesterol concentration were significantly higher in May then in June (20,55 L/d vs. 15,44 L/d; 4,55 mmol/L vs. 3,96 mmol/L, respectively, whereas the GPx and GGT activity was significantly lower in May in comparison with June (GPx=1720,66 U/L vs. GPx=1808 U/L, and GGT = 20,11 U/L vs. 23,22 U/L, respectively. A positive correlation between the milk production and the total blood cholesterol level was detected (r=0,58. The total blood cholesterol concentration in the blood plasma mighthave been one of the indicators of production performance in the Simmental cows herd, whereas the activity of stated enzymes may specify the nutritional status thereof the milk production depends on.

  3. Reliability of the Dinamap non-invasive monitor in the measurement of blood pressure of ill Asian newborns.

    Science.gov (United States)

    Chia, F; Ang, A T; Wong, T W; Tan, K W; Fung, K P; Lee, J; Khin, K

    1990-05-01

    Four hundred thirty-one paired sets of readings of systolic and diastolic blood pressure and 438 paired sets of readings of mean arterial BP from 49 ill newborns, including 21 very low birth weight infants, were analyzed for the extent and pattern of agreement and the linear relationship between the Dinamap oscillometric monitor and the direct intraarterial blood pressure readings. Agreement between the two methods was measured by the intraclass correlation, whereas the linear relationship was assessed by the product-moment correlation. The intraclass correlations for systolic, diastolic and mean blood pressures were 0.696, 0.766, and 0.781, respectively. The product-moment correlations for systolic, diastolic and mean blood pressures were 0.706, 0.768, and 0.786, respectively. BP measurements by the Dinamap monitor showed reasonably close agreement to those obtained by the intraarterial mean arterial pressure ranges above 40 mmHg. For mean arterial pressure of 40 mmHg and lower, BP readings by the Dinamap monitor tended to be higher than those obtained by the intraarterial method. These findings appeared to be consistent regardless of the birth weight of the newborn. PMID:2340688

  4. The Blood Quality Monitoring Network System's Role in the Safety of Blood Transfusion%血液质量监测网络体系在输血安全中的作用

    Institute of Scientific and Technical Information of China (English)

    胡秀兰; 田志彬; 李忠杰; 玄红霞; 陈书芳

    2013-01-01

    目的探寻建立血站血液质量监测网络体系,确保血液质量与安全。方法对献血者献血前的血液标本进行初筛,对采血场所、关键原辅材料、关键设备进行质量控制,同时加强采后血液的质量监测以及各供血库的监测。结果比较质量监测网络体系建立前后血液质量情况,发现血液质量监测网络体系建立后血液质量有明显提高。结论建立完整有效、安全的血液质量监测网络体系,对血液的采集、检验、保存、运送和使用等多个环节进行血液质量监测,监控血液"从血管到血管"的全过程,确保献血者、用血者身体健康和生命安全。%Objective To explore the establishment of a blood bank blood quality monitoring network system to ensure the quality and safety of blood. Methods for screening blood donors before blood samples, blood col ection establishments, the key raw materials, key equipment for quality control quality, while strengthening the postharvest quality monitoring of blood as wel as monitoring for blood banks. Results the comparative quality monitoring network system to establish the quality of blood before and after situation, significantly improved blood quality monitoring network system to establish the quality of blood. Conclusions to establish a complete and ef ective, safe blood quality monitoring network system for blood col ection, testing, preservation, transportation and use of other aspects of the blood quality monitoring, monitoring of blood"from the blood vessels to the blood vessels,"the whole process to ensure donors with blood by the health and safety of life.

  5. Whole blood is the sample matrix of choice for monitoring systemic triclocarban levels.

    Science.gov (United States)

    Schebb, Nils Helge; Ahn, Ki Chang; Dong, Hua; Gee, Shirley J; Hammock, Bruce D

    2012-05-01

    The antibacterial triclocarban (TCC) concentrates in the cellular fraction of blood. Consequently, plasma levels are at least two-fold lower than the TCC amount present in blood. Utilizing whole blood sampling, a low but significant absorption of TCC from soap during showering is demonstrated for a small group of human subjects. PMID:22273184

  6. Whole blood is the sample matrix of choice for monitoring systemic triclocarban levels

    OpenAIRE

    Schebb, Nils Helge; Ahn, Ki Chang; Dong, Hua; Gee, Shirley J.; Hammock, Bruce D.

    2012-01-01

    The antibacterial triclocarban (TCC) concentrates in the cellular fraction of blood. Consequently, plasma levels are at least two-fold lower than the TCC amount present in blood. Utilizing whole blood sampling, a low but significant absorption of TCC from soap during showering is demonstrated for a small group of human subjects.

  7. Intraoperative laser speckle contrast imaging for monitoring cerebral blood flow: results from a 10-patient pilot study

    Science.gov (United States)

    Richards, Lisa M.; Weber, Erica L.; Parthasarathy, Ashwin B.; Kappeler, Kaelyn L.; Fox, Douglas J.; Dunn, Andrew K.

    2012-02-01

    Monitoring cerebral blood flow (CBF) during neurosurgery can provide important physiological information for a variety of surgical procedures. Although multiple intraoperative vascular monitoring technologies are currently available, a quantitative method that allows for continuous monitoring is still needed. Laser speckle contrast imaging (LSCI) is an optical imaging method with high spatial and temporal resolution that has been widely used to image CBF in animal models in vivo. In this pilot clinical study, we adapted a Zeiss OPMI Pentero neurosurgical microscope to obtain LSCI images by attaching a camera and a laser diode. This LSCI adapted instrument has been used to acquire full field flow images from 10 patients during tumor resection procedures. The patient's ECG was recorded during acquisition and image registration was performed in post-processing to account for pulsatile motion artifacts. Digital photographs confirmed alignment of vasculature and flow images in four cases, and a relative change in blood flow was observed in two patients after bipolar cautery. The LSCI adapted instrument has the capability to produce real-time, full field CBF image maps with excellent spatial resolution and minimal intervention to the surgical procedure. Results from this study demonstrate the feasibility of using LSCI to monitor blood flow during neurosurgery.

  8. Evaluation of OneTouch Verio(®), a new blood glucose self-monitoring system for patients with diabetes

    DEFF Research Database (Denmark)

    Littmann, Karin; Petersen, Eva; Pussinen, Christel;

    2013-01-01

    Introduction. Self-monitoring of blood glucose (SMBG) is important in diabetes management. Reliable and user-friendly instruments are essential. OneTouch Verio(®) is a new blood glucose concentration-measuring system designed to be used by patients with diabetes and healthcare professionals. The ...... between strip lots. Conclusion. The OneTouch Verio(®) displayed sufficient analytical quality and satisfactory user-friendliness. It is suitable for point-of-care testing of blood glucose concentration when handled by patients and healthcare professionals.......Introduction. Self-monitoring of blood glucose (SMBG) is important in diabetes management. Reliable and user-friendly instruments are essential. OneTouch Verio(®) is a new blood glucose concentration-measuring system designed to be used by patients with diabetes and healthcare professionals....... The objective of the present study was to evaluate the analytical performance of the OneTouch Verio(®). Method. The OneTouch Verio(®) was evaluated by the Scandinavian evaluation of laboratory equipment for primary healthcare (SKUP) according to a protocol based on ISO 15197 and the American Diabetes...

  9. Evaluation of OneTouch Verio, a new blood glucose self-monitoring system for patients with diabetes

    DEFF Research Database (Denmark)

    Littman, Karin; Petersen, Eva R.B.; Pussinen, Christel;

    2013-01-01

    Introduction. Self-monitoring of blood glucose (SMBG) is important in diabetes management. Reliable and user-friendly instruments are essential. OneTouch Verio® is a new blood glucose concentration-measuring system designed to be used by patients with diabetes and healthcare professionals. The ob....... Conclusion. The OneTouch Verio® displayed sufficient analytical quality and satisfactory user-friendliness. It is suitable for point-of-care testing of blood glucose concentration when handled by patients and healthcare professionals.......Introduction. Self-monitoring of blood glucose (SMBG) is important in diabetes management. Reliable and user-friendly instruments are essential. OneTouch Verio® is a new blood glucose concentration-measuring system designed to be used by patients with diabetes and healthcare professionals....... The objective of the present study was to evaluate the analytical performance of the OneTouch Verio®. Method. The OneTouch Verio® was evaluated by the Scandinavian evaluation of laboratory equipment for primary healthcare (SKUP) according to a protocol based on ISO 15197 and the American Diabetes Association...

  10. Therapeutic drug monitoring in dried blood spots using liquid microjunction surface sampling and high resolution mass spectrometry.

    Science.gov (United States)

    Gaissmaier, Tanja; Siebenhaar, Markus; Todorova, Vanya; Hüllen, Volker; Hopf, Carsten

    2016-02-01

    Dried blood spots (DBS) are a versatile and stable tool for direct clinical blood analysis. Ambient high-resolution mass spectrometry is emerging as a method of choice for their quantitative analysis, for instance in therapeutic drug monitoring. Here, we coupled liquid microjunction surface sampling technology, a so-called Flowprobe, with an Orbitrap mass spectrometer and demonstrated the utility of this set-up for direct quantification of multiple drugs in DBS on filter paper. A three-layer set-up that we had introduced earlier enabled introduction of internal standards into DBS. We furthermore took an established point-of-care test system a step further and analyzed disposable test fields for blood glucose monitoring also for Flowprobe-based acetaminophen screening without additional sample preparation. Using as little as 2 μL blood, the method had an LOD of 1 μg mL(-1) (coefficient of variation of ≤15%) and acetaminophen recoveries of 82 to 119% for blinded samples, as assessed by LC-MS/MS. Half an hour after ingestions of a single 1000 mg acetaminophen dose, indistinguishable drug levels were measured in three healthy volunteers by LC-MS/MS and Flowprobe-Orbitrap MS analysis of DBS. Flowprobe analysis of DBS was 6- to 100-times more sensitive than corresponding desorption electrospray ionization MS analysis for four drugs. For instance, the LOD for salicylic acid analysis was 0.07 ng mL(-1) with Flowprobe measurement. Furthermore, we showed that multi-component analysis of five different substances, which may mimic polypharmacy in diabetes patients, in one blood sample for screening purposes was feasible. Taken together, our study suggests that microjunction surface sampling of DBS on filter paper and disposable point-of-care test fields may be developed into routine methods for near-patient multi-compound therapeutic drug monitoring that may advance blood screening analysis for patients with polypharmacy. PMID:26759833

  11. Gas Bubble Disease Monitoring and Research of Juvenile Salmonids : Annual Report 1996.

    Energy Technology Data Exchange (ETDEWEB)

    Maule, Alec G.; Beeman, John W.; Hans, Karen M.; Mesa, M.G.; Haner, P.; Warren, J.J. [Geological Survey, Cook, WA (United States). Columbia River Research Lab.

    1997-10-01

    This document describes the project activities 1996--1997 contract year. This report is composed of three chapters which contain data and analyses of the three main elements of the project: field research to determine the vertical distribution of migrating juvenile salmonids, monitoring of juvenile migrants at dams on the Snake and Columbia rivers, and laboratory experiments to describe the progression of gas bubble disease signs leading to mortality. The major findings described in this report are: A miniature pressure-sensitive radio transmitter was found to be accurate and precise and, after compensation for water temperature, can be used to determine the depth of tagged-fish to within 0.32 m of the true depth (Chapter 1). Preliminary data from very few fish suggest that depth protects migrating juvenile steelhead from total dissolved gas supersaturation (Chapter 1). As in 1995, few fish had any signs of gas bubble disease, but it appeared that prevalence and severity increased as fish migrated downstream and in response to changing gas supersaturation (Chapter 2). It appeared to gas bubble disease was not a threat to migrating juvenile salmonids when total dissolved gas supersaturation was < 120% (Chapter 2). Laboratory studies suggest that external examinations are appropriate for determining the severity of gas bubble disease in juvenile salmonids (Chapter 3). The authors developed a new method for examining gill arches for intravascular bubbles by clamping the ventral aorta to reduce bleeding when arches were removed (Chapter 3). Despite an outbreak of bacterial kidney disease in the experimental fish, the data indicate that gas bubble disease is a progressive trauma that can be monitored (Chapter 3).

  12. Ambulatory blood pressure monitoring after 1 year on valsartan or amlodipine-based treatment: a VALUE substudy

    DEFF Research Database (Denmark)

    Pedersen, Ole Lederballe; Mancia, Giuseppe; Pickering, Thomas;

    2007-01-01

    OBJECTIVE: The ambulatory blood pressure (ABP) monitoring substudy of the Valsartan Antihypertensive Long-term Use Evaluation (VALUE) trial was carried out in a subset of patients from USA, Italy and Denmark. ABP was measured after 1 year in the trial, with the aim of evaluating comparability...... of ABP levels on valsartan (VAL) and amlodipine (AML)-based regimens. METHODS: ABP was measured every 20 min during a 25-h period after morning administration of medicine; 659 patients were available for intention-to-treat analysis. RESULTS: Office blood pressure (BP) differences were smaller than...

  13. Design and Evaluation of a Fully Implantable Control Unit for Blood Pumps

    Directory of Open Access Journals (Sweden)

    Kristin Unthan

    2015-01-01

    Full Text Available As the number of donor hearts is limited while more and more patients suffer from end stage biventricular heart failure, Total Artificial Hearts become a promising alternative to conventional treatment. While pneumatic devices sufficiently supply the patients with blood flow, the patient’s quality of life is limited by the percutaneous pressure lines and the size of the external control unit. This paper describes the development of the control unit of the ReinHeart, a fully implantable Total Artificial Heart. General requirements for any implantable control unit are defined from a technical and medical point of view: necessity of a Transcutaneous Energy Transmission, autonomous operation, safety, geometry, and efficiency. Based on the requirements, a prototype is designed; it incorporates a LiFePo4 battery pack with charger, a rectifier for transcutaneous energy transmission, the motor’s driver electronics, and a microcontroller which monitors and controls all functions. In validation tests, the control unit demonstrated a stable operation on TET and battery supply and a safe switching from one supply to the other. The overall mean efficiency is 14% on TET and 22% on battery supply. The control unit is suitable for chronic animal trials of the ReinHeart.

  14. Body Position Influences Which Neural Structures Are Recruited by Lumbar Transcutaneous Spinal Cord Stimulation

    OpenAIRE

    Danner, Simon M.; Matthias Krenn; Hofstoetter, Ursula S.; Andrea Toth; Winfried Mayr; Karen Minassian

    2016-01-01

    Transcutaneous stimulation of the human lumbosacral spinal cord is used to evoke spinal reflexes and to neuromodulate altered sensorimotor function following spinal cord injury. Both applications require the reliable stimulation of afferent posterior root fibers. Yet under certain circumstances, efferent anterior root fibers can be co-activated. We hypothesized that body position influences the preferential stimulation of sensory or motor fibers. Stimulus-triggered responses to transcutaneous...

  15. Clinical study of transcutaneous vaccination using a hydrogel patch for tetanus and diphtheria.

    Science.gov (United States)

    Hirobe, Sachiko; Matsuo, Kazuhiko; Quan, Ying-Shu; Kamiyama, Fumio; Morito, Hironori; Asada, Hideo; Takaya, Yusuke; Mukai, Yohei; Okada, Naoki; Nakagawa, Shinsaku

    2012-02-27

    Transcutaneous immunization (TCI) is a non-invasive and easy-to-use vaccination method. We demonstrated the efficacy and safety of a transcutaneous vaccine formulation using a hydrogel patch in animal experiments. In the present study, we performed a clinical study to apply our TCI formulation for vaccination against tetanus and diphtheria in human. The TCI device was a hydrogel patch (antigen-free) applied to the left brachial medial skin of 22 healthy volunteers for 48 h. Next, the hydrogel patch, containing 2mg tetanus toxoid (TT) and 2mg diphtheria toxoid (DT) as the TCI formulation, was applied to 27 healthy volunteers for 24h and some volunteers were vaccinated again by TCI formulation. For safety assessment, the patch application site was observed to assess local adverse events, and systemic adverse events were determined by a blood test. The antigen-free hydrogel patch and TCI formulation containing TT and DT did not induce local or systemic severe adverse events. For vaccine efficacy estimation, toxoid-specific serum antibody titers were determined by ELISA and the toxin-neutralizing activity of the induced antibody was evaluated in a passive-challenge experiment. The anti-TT IgG titer and the anti-DT IgG titer increased, and a significant effect was detected by paired t-test. The antibody titers were maintained at higher level than that before vaccination for at least 1 year. Moreover, toxoid-specific antibodies were produced by the second vaccination in some subjects. Antibodies induced by application of the TCI formulation neutralized the toxin and prevented toxic death in mice. In addition, changes in the skin condition due to application of the TCI formulation were observed under in vivo confocal Raman spectroscopy. The amount of water and patch components in the stratum corneum increased after application of the TCI formulation, suggesting that the change in the skin condition was related to antigen penetration. These data indicate that this easy

  16. Loading of red blood cells with an analyte-sensitive dye for development of a long-term monitoring technique

    Science.gov (United States)

    Ritter, Sarah C.; Meissner, Kenith E.

    2012-03-01

    Measurement of blood analytes, such as pH and glucose, provide crucial information about a patient's health. Some such analytes, such as glucose in the case of diabetes, require long-term or near-continuous monitoring for proper disease management. However, current monitoring techniques are far from ideal: multiple-per-day finger stick tests are inconvenient and painful for the patient; implantable sensors have short functional life spans (i.e., 3-7 days). Red blood cells serve as an attractive alternative for carriers of analyte sensors. Once reintroduced to the blood stream, these carriers may continue to live for the remainder of their life span (120 days for humans). They are also biodegradable and biocompatible, thereby eliminating the immune system response common for many implanted devices. The proposed carrier system takes advantage of the ability of the red blood cells to swell in response to a decrease in the osmolarity of the extracellular solution. Just before the membranes lyse, they develop small pores on the scale of tens of nanometers. Analyte-sensitive dyes in the extracellular solution may then diffuse into the perforated red blood cells and become entrapped upon restoration of physiological temperature and osmolarity. Because the membranes contain various analyte transporters, intracellular analyte levels rapidly equilibrate to those of the extracellular solution. A fluorescent dye has been loaded inside of red blood cells using a preswelling technique. Alterations in preparation parameters have been shown to affect characteristics of the resulting dye-loaded red blood cells (e.g., intensity of fluorescence).

  17. A Gas-Jet Profile Monitor for the CLIC Drive Beam

    CERN Document Server

    Jeff, A; Lefevre, T; Tzoganis, V; Welsch, C P

    2013-01-01

    The Compact Linear Collider (CLIC) will use a novel acceleration scheme in which energy extracted from a very intense beam of relatively low-energy electrons (the Drive Beam) is used to accelerate a lower intensity Main Beam to very high energy. The high intensity of the Drive Beam, with pulses of more than 1015 electrons, poses a challenge for conventional profile measurements such as wire scanners. Thus, new non-invasive profile measurements are being investigated. Profile monitors using gas ionisation or fluorescence have been used at a number of accelerators. Typically, extra gas must be injected at the monitor and the rise in pressure spreads for some distance down the beam pipe. In contrast, a gas jet can be fired across the beam into a receiving chamber, with little gas escaping into the rest of the beam pipe. In addition, a gas jet shaped into a thin plane can be used like a screen on which the beam crosssectionis imaged. In this paper we present some arrangements for the generation of such a jet. In ...

  18. Continued Development of Compact Multi-gas Monitor for Life Support Systems Control in Space

    Science.gov (United States)

    Delgado-Alonso, Jesús; Phillips, Straun; Chullen, Cinda; Quinn, Gregory

    2016-01-01

    Miniature optic gas sensors (MOGS) based on luminescent materials have shown great potential as alternatives to Near-Infrared-based gas sensor systems for the advanced space suit portable life support system (PLSS). The unique capability of MOGS for carbon dioxide and oxygen monitoring under wet conditions has been reported, as has the fast recovery of MOGS humidity sensors after long periods of being wet. Lower volume and power requirements are also potential advantages of MOGS over both traditional and advanced Non-Dispersive Infrared (NDIR) gas sensors, which have shown so far longer life than luminescent sensors. This paper presents the most recent results in the development and analytical validation of a compact multi-gas sensor unit based on luminescent sensors for the PLSS. Results of extensive testing are presented, including studies conducted at Intelligent Optical Systems laboratories, a United Technology Corporation Aerospace Systems (UTAS) laboratory, and a Johnson Space Center laboratory. The potential of this sensor technology for gas monitoring in PLSSs and other life support systems and the advantages and limitations found through detailed sensor validation are discussed.

  19. Electrochemical sensor for continuous transcutaneous PCO2 measurement.

    Science.gov (United States)

    Beran, A V; Huxtable, R F; Sperling, D R

    1976-09-01

    A sensor suitable for continuous transcutaneous PCO2 measurements is described. The sensor consists of an antimony-antimony oxide electrode in combination with a silver-silver chloride reference electrode, bathed in an electrolyte and covered by a Teflon membrane. A servo-controlled heater unit was used to maintain the sensor's temperature and to produce local hyperemia. The resulting oxidation-reduction potential under constant temperature is a linear function of the logarithm PCO2. Response time (95%) to step changes in PCO2 from 27 to 70 mmHg was 2.7 +/- 0.3 min. Following a 12-h "aging" time, the electrode exhibited a minimal drift of 5.2 +/- 2.2 mV for 16 h, representing an average PCO2 drift of 0.5 mmHg/h. This sensor was applied on three rabbits and on five human volunteers, and found satisfactory under normal physiological conditions.

  20. [Multicenter evaluation of the reliability of five blood glucose monitoring systems].

    Science.gov (United States)

    Jday-Daly, Ismahen; Augereau-Vacher, Christine; De Curraize, Claire; Fonfrède, Michèle; Lefevre, Guillaume; Lacour, Bernard; Hennequin-Le Meur, Carole

    2011-01-01

    As part of a tender AP-HP Paris Hospitals, an assessment of the reliability record of five blood glucose monitoring systems (BGMSs) (Optium Xceed (Abbott), Contour TS (Bayer), One Touch Ultra (Lifescan), Stat Strip Xpress (Nova) and Accu Check (Roche) and an evaluation of their sensitivity to changes in hematocrit were conducted in 4 hospitals of Paris. In terms of inaccuracy, all BGMSs have submitted CV repetability under the limits of acceptability. One BGMS (Lifescan) presented a CV of reproducibility outside limit of acceptability (13.1%). The inaccuracy was measured by a comparison method on multiparameter analyser relative to the hexokinase method for two sites, the glucose oxidase for the two others. The coefficients of correlation varied from 0.8405 to 0.9303. However, according to both defined acceptability criteria (absolute value difference between the result acquired on analyzer and those determined with the BGMS), the percentage of results outside acceptability was above 20% for two BGMSs (Abbott and Lifescan). Similarly, a net effect of changes in hematocrit was observed on the results of those two BGMSs. BGMS Nova was the most reliable, because of the correction device for hematocrit and blank substractions owed to interferences. In terms of expertise, BGMSs Nova and Roche have been selected with the best analytical performance and practicability satisfactory. In the future, accreditation with standard NF/EN 22870 requested for point of care testing, will require a close collaboration between biologists and clinicians to establish a system of strict quality control to detect deviations of these BGMSs. PMID:21463996

  1. Controlled ultrasound-induced blood-brain barrier disruption using passive acoustic emissions monitoring.

    Directory of Open Access Journals (Sweden)

    Costas D Arvanitis

    Full Text Available The ability of ultrasonically-induced oscillations of circulating microbubbles to permeabilize vascular barriers such as the blood-brain barrier (BBB holds great promise for noninvasive targeted drug delivery. A major issue has been a lack of control over the procedure to ensure both safe and effective treatment. Here, we evaluated the use of passively-recorded acoustic emissions as a means to achieve this control. An acoustic emissions monitoring system was constructed and integrated into a clinical transcranial MRI-guided focused ultrasound system. Recordings were analyzed using a spectroscopic method that isolates the acoustic emissions caused by the microbubbles during sonication. This analysis characterized and quantified harmonic oscillations that occur when the BBB is disrupted, and broadband emissions that occur when tissue damage occurs. After validating the system's performance in pilot studies that explored a wide range of exposure levels, the measurements were used to control the ultrasound exposure level during transcranial sonications at 104 volumes over 22 weekly sessions in four macaques. We found that increasing the exposure level until a large harmonic emissions signal was observed was an effective means to ensure BBB disruption without broadband emissions. We had a success rate of 96% in inducing BBB disruption as measured by in contrast-enhanced MRI, and we detected broadband emissions in less than 0.2% of the applied bursts. The magnitude of the harmonic emissions signals was significantly (P<0.001 larger for sonications where BBB disruption was detected, and it correlated with BBB permeabilization as indicated by the magnitude of the MRI signal enhancement after MRI contrast administration (R(2 = 0.78. Overall, the results indicate that harmonic emissions can be a used to control focused ultrasound-induced BBB disruption. These results are promising for clinical translation of this technology.

  2. Novel industrial application: flammable and toxic gas monitoring in the printing industry

    Science.gov (United States)

    Jacobson, Esther; Spector, Yechiel

    1999-12-01

    The present paper describes an Open Path Electro-Optical Gas Monitoring System specifically designed for in-situ on-line monitoring of flammable and toxic atmospheres in the Printing Industry in general, and for air-duct applications in particular. The printing industry posies unique fire hazards due to the variety of toxic and flammable chemical employed in the various printing process. Flammable material such as paper, ink, solvents, thinners, metal powders, cornstarch powders, cloth, synthetic materials are frequently used in the printing industry in several processes such as letter-pressing, lithography, screen printing etc.

  3. Implementation of the ABL-90 blood gas analyzer in a ground-based mobile emergency care unit

    DEFF Research Database (Denmark)

    Mikkelsen, Søren; Wolsing-Hansen, Jonathan; Nybo, Mads;

    2015-01-01

    Point-of Care analysis is increasingly being applied in the prehospital scene. Arterial blood gas analysis is one of many new initiatives adding to the diagnostic tools of the prehospital physician. In this paper we present a study on the feasibility of the Radiometer ABL-90 in a ground...

  4. Reliability of point-of-care hematocrit, blood gas, electrolyte, lactate and glucose measurement during cardiopulmonary bypass.

    NARCIS (Netherlands)

    Steinfelder-Visscher, J.; Weerwind, P.W.; Teerenstra, S.; Brouwer, M.H.J.

    2006-01-01

    BACKGROUND: Recently, the GEM Premier blood gas analyser was upgraded to the GEM Premier 3000. In addition to pH, pCO2, pO2, Na+, K+, Ca2+, and hematocrit measurement, glucose and lactate can be measured on the GEM Premier 3000. In this prospective clinical study, the analytical performance of the G

  5. Stratification of Ambulatory Blood Pressure Monitoring Findings by Cluster Analysis in Patients with Arterial Hypertension, Obesity and Albuminuria

    OpenAIRE

    Samoyavcheva S.V.; Shkarin Vl.V.

    2013-01-01

    The aim of the investigation was to study the characteristics of ambulatory blood pressure monitoring (ABPM) indices in the combination of arterial hypertension (AH) with obesity and albuminuria using cluster analysis. Material and Methods. The study involved 70 AH patients randomly chosen, aged from 23 to 71 years (mean age — 47.9 years). ABPM was performed before antihypertensive therapy administration. We estimated body mass index and albuminuria level. ABPM indices were stratified int...

  6. The influence of self-owned home blood pressure monitoring (HBPM) on primary care patients with hypertension: A qualitative study

    OpenAIRE

    Abdullah Adina; Othman Sajaratulnisah

    2011-01-01

    Abstract Background Home blood pressure monitoring (HBPM) is gaining popularity among hypertensive patients. This study aimed to explore the influence of self-initiated HBPM on primary care patients with hypertension. Methods Six in-depth interviews and two focus group discussions were conducted, taking into consideration the experiences of 24 primary care patients with hypertension. These patients had been using HBPM as part of their hypertension management. The overriding influences were gr...

  7. 8th Annual Symposium on Self-Monitoring of Blood Glucose (SMBG): April 16-18, 2015, Republic of Malta.

    Science.gov (United States)

    Parkin, Christopher G; Homberg, Anita; Hinzmann, Rolf

    2015-11-01

    International experts in the fields of diabetes, diabetes technology, endocrinology, mobile health, sport science, and regulatory issues gathered for the 8(th) Annual Symposium on Self-Monitoring of Blood Glucose (SMBG) with a focus on personalized diabetes management. The aim of this meeting was to facilitate new collaborations and research projects to improve the lives of people with diabetes. The 2015 meeting comprised a comprehensive scientific program, parallel interactive workshops, and two keynote lectures.

  8. A novel artificial anal sphincter system based on transcutaneous energy transmission

    Institute of Scientific and Technical Information of China (English)

    Zan Peng; Yan Guozheng; Liu Hua

    2008-01-01

    For controlling anal incontinence, a new artificial anal sphincter system (AASS) with sensor feedback based on transcutaneous energy transmission is developed. The device mainly comprises an artificial anal sphincter (AAS), a wireless power supply subsystem, and a communication subsystem. The artificial anal sphincter comprises a front cuff and a sensor cuff placed around the rectum, a reservoir sited in abdominal cavity and a micropump controlling inflation and deflation of the front cuff. There are two pressure sensors in the artificial anal sphincter. One can measure the pressure in the front cuff to clamp the rectum, the other in the sensor cuff can measure the pressure of the rectum. Wireless power supply subsystem includes a resonance transmit coil to transmit an alternating magnetic field and a secondary coil to receive the power. Wireless communication subsystem can transmit the pressure information of the artificial anal sphincter to the monitor, or send the control commands to the artificial anal sphincter. A prototype is designed and the basic function of the artificial anal sphincter system has been tested through experiments. The results demonstrate that the artificial anal sphincter system can control anal incontinence effectively.

  9. Optical Multi-Gas Monitor Technology Demonstration on the International Space Station

    Science.gov (United States)

    Pilgrim, Jeffrey S.; Wood, William R.; Casias, Miguel E.; Vakhtin, Andrei B,; Johnson, Michael D.; Mudgett, Paul D.

    2014-01-01

    There are a variety of both portable and fixed gas monitors onboard the International Space Station (ISS). Devices range from rack-mounted mass spectrometers to hand-held electrochemical sensors. An optical Multi-Gas Monitor has been developed as an ISS Technology Demonstration to evaluate long-term continuous measurement of 4 gases. Based on tunable diode laser spectroscopy, this technology offers unprecedented selectivity, concentration range, precision, and calibration stability. The monitor utilizes the combination of high performance laser absorption spectroscopy with a rugged optical path length enhancement cell that is nearly impossible to misalign. The enhancement cell serves simultaneously as the measurement sampling cell for multiple laser channels operating within a common measurement volume. Four laser diode based detection channels allow quantitative determination of ISS cabin concentrations of water vapor (humidity), carbon dioxide, ammonia and oxygen. Each channel utilizes a separate vertical cavity surface emitting laser (VCSEL) at a different wavelength. In addition to measuring major air constituents in their relevant ranges, the multiple gas monitor provides real time quantitative gaseous ammonia measurements between 5 and 20,000 parts-per-million (ppm). A small ventilation fan draws air with no pumps or valves into the enclosure in which analysis occurs. Power draw is only about 3 W from USB sources when installed in Nanoracks or when connected to 28V source from any EXPRESS rack interface. Internal battery power can run the sensor for over 20 hours during portable operation. The sensor is controlled digitally with an FPGA/microcontroller architecture that stores data internally while displaying running average measurements on an LCD screen and interfacing with the rack or laptop via USB. Design, construction and certification of the Multi-Gas Monitor were a joint effort between Vista Photonics, Nanoracks and NASA-Johnson Space Center (JSC

  10. Development of a Flexible Implantable Sensor for Postoperative Monitoring of Blood Flow

    OpenAIRE

    Cannata, Jonathan M.; Chilipka, Thomas; Yang, Hao-Chung; Han, Sukgu; Ham, Sung W.; Rowe, Vincent L.; Weaver, Fred A; Shung, K. Kirk; Vilkomerson, David

    2012-01-01

    We have developed a blood flow measurement system using Doppler ultrasound flow sensors fabricated of thin and flexible piezoelectric-polymer films. These flow sensors can be wrapped around a blood vessel and accurately measure flow. The innovation that makes this flow sensor possible is the diffraction-grating transducer. A conventional transducer produces a sound beam perpendicular to its face; therefore, when placed on the wall of a blood vessel, the Doppler shift in the backscattered ultr...

  11. Self-monitoring of tear glucose: the development of a tear based glucose sensor as an alternative to self-monitoring of blood glucose.

    Science.gov (United States)

    La Belle, Jeffrey T; Adams, Anngela; Lin, Chi-En; Engelschall, Erica; Pratt, Breanna; Cook, Curtiss B

    2016-07-28

    Tear glucose sensing for diabetes management has long been sought as an alternative to more invasive self-monitoring of blood glucose (SMBG). However, tear glucose sensors were known to have limitations, including correlation issues with blood glucose due to low sample volume, low concentration of glucose in the tear fluid, and evaporation of the tear sample. An engineering design approach to solve these problems led to the development of an integrated device capable of collecting the tear sample from the ocular surface with little to no stress on the eye, with an extremely low limit of detection, broad dynamic range, and rapid detection and analysis of sample. Here we present the development of a prototypical self-monitoring of tear glucose (SMTG) sensor, summarizing bench studies on the enzymes and their specificity, the development of the fluid capture device and its manufacture and performance and results of system testing in an animal study where safety, lag time and tear glucose to blood glucose correlation were assessed. PMID:27327531

  12. A plunger lift and monitoring system for gas wells based on deployment-retrievement integration

    Directory of Open Access Journals (Sweden)

    Zheng Tong

    2015-11-01

    Full Text Available As a necessary step, removing liquid in the wellbore plays an important role during the production of gas wells. Plunger lift is a widely-used intermittent deliquification process for gas wells. However, the manual control way and wire logging are still utilized as a downhole monitoring way for plunger lift, which is not efficient in terms of interrupting the production. This paper presents an improved solution that logging instruments canister are deployed and retrieved by means of a new assembly. With the reciprocating plunger, logging instruments canister can be carried and deployed to the bottom of a gas well to carry out logging and sampling tasks on the production demand of a field. After the deployment and logging tasks are performed, logging instruments canister is carried back to the surface by the plunger and then data is transferred to the wellhead device near field wireless communication technology. This newly developed plunger lift system comprises plunger body, deployment sub-assembly, retrieve sub-assembly and logging instruments canister. The surface device comprises RF antenna, reader and writer. Based upon the method of deployment-retrieve integration, the new deliquification process is introduced and on-line monitoring of production dynamics can be performed including P/T measurement, downhole fluid sampling, pressure build-up, etc. without interrupting production. The general solution and engineering design parameters have been confirmed by research teams, while system prototype manufacture and workbench tests are being performed. The cost-effective way combining deliquification with dynamic monitoring is developed and contributes to increasing production and the stable productivity of gas wells. It is very significant for low-pressure and low-production gas fields to achieve automation production and management.

  13. Clinical significance of blood gas and electrolyte analysis, CK, CK-MB and HBDH changes in neonatal asphyxia

    Institute of Scientific and Technical Information of China (English)

    Bao-Hua Xu; Xin Lin; Mi-Jia Huang

    2016-01-01

    Objective:To investigate the clinical significance of blood gas and electrolyte analysis, CK, CK-MB and HBDH changes in neonatal asphyxia.Methods:A total of 100 newborns with asphyxia who visited in our hospital were collected, and divided into severe group (n=20) and mild group (n=80) according to the asphyxia degree, and 50 healthy newborns regarded as control group. The 3 groups received blood gas analysis (pH, BE and PaCO2), electrolyte (K+, Na+ and Ca2+) and 3 kinds of enzymes (CK, CK-MB and HBDH) were tested and compared.Results: Compared with control group, pH and BE of blood gas indexes decreased significantly and PaCO2 increased significantly in severe group (P0.05). Compared with mild group, pH and BE of blood gas indexes decreased significantly and PaCO2 increased significantly in severe group (P0.05). Compared with mild group, the level of Ca2+ decreased significantly in severe group (P<0.05); Compared with control group, the levels of CK, CK-MB and HBDH increased significantly in severe and mild group (P<0.05). Compared with mild group, the levels of CK, CK-MB and HBDH increased significantly in severe group (P<0.05).Conclusions:The detection of blood gas and electrolyte analysis, CK, CK-MB and HBDH can provide an objective evidence for the diagnosis of neonatal asphyxia and estimation of severity degree which was helpful in clinical treatment.

  14. Optical Multi-Gas Monitor Technology Demonstration on the International Space Station

    Science.gov (United States)

    Pilgrim, Jeffrey S.; Wood, William R.; Casias, Miguel E.; Vakhtin, Andrei B.; Johnson, Michael D.; Mudgett, Paul D.

    2014-01-01

    The International Space Station (ISS) employs a suite of portable and permanently located gas monitors to insure crew health and safety. These sensors are tasked with functions ranging from fixed mass spectrometer based major constituents analysis to portable electrochemical sensor based combustion product monitoring. An all optical multigas sensor is being developed that can provide the specificity of a mass spectrometer with the portability of an electrochemical cell. The technology, developed under the Small Business Innovation Research program, allows for an architecture that is rugged, compact and low power. A four gas version called the Multi-Gas Monitor was launched to ISS in November 2013 aboard Soyuz and activated in February 2014. The portable instrument is comprised of a major constituents analyzer (water vapor, carbon dioxide, oxygen) and high dynamic range real-time ammonia sensor. All species are sensed inside the same enhanced path length optical cell with a separate vertical cavity surface emitting laser (VCSEL) targeted at each species. The prototype is controlled digitally with a field-programmable gate array/microcontroller architecture. The optical and electronic approaches are designed for scalability and future versions could add three important acid gases and carbon monoxide combustion product gases to the four species already sensed. Results obtained to date from the technology demonstration on ISS are presented and discussed.

  15. Tests on a prototype of the Passive Fission Gas Monitor for failed detection (PRISM reactor)

    International Nuclear Information System (INIS)

    The Passive Diffusion Fission Gas Monitor PDFGM is mounted on the PRISM reactor head and extends into the cover gas Region where it determines the presence of radioactive fission gases (Kr, Xe, and so on) released from failed fuel pins. It contains a steel diffusion column that is closed at the upper end but opened to the cover gas at its lower end. The upper portion of the column is located in the field of view of a collimated gamma detector which is shielded from the remainder of the cover gas and of the sodium pool below. Heaters are provided to obtain a uniform axial temperature in the gas column and to minimize the potential for natural convection currents. In this way, the molecular diffusion can be established based on the fission gas concentration gradients along the column length. This is an advanced solution in comparison with current devices based on active components (pumps, filters, and so on). The experimental results on a prototype of PDFGM and their interpretation will be presented in this paper. (author)

  16. Calculation and verification of blood ethanol measurement uncertainty for headspace gas chromatography.

    Science.gov (United States)

    Sklerov, Jason H; Couper, Fiona J

    2011-09-01

    An estimate was made of the measurement uncertainty for blood ethanol testing by headspace gas chromatography. While uncertainty often focuses on compliance to a single threshold level (0.08 g/100 mL), the existence of multiple thresholds, related to enhanced sentencing, subject age, or commercial vehicle licensure, necessitate the use of an estimate with validity across multiple specification levels. The uncertainty sources, in order of decreasing magnitude, were method reproducibility, linear calibration, recovery, calibrator preparation, reference material, and sample preparation. A large set of reproducibility data was evaluated (n = 15,433) in order to encompass measurement variability across multiple conditions, operators, instruments, concentrations and timeframes. The relative, combined standard uncertainty was calculated as ±2.7%, with an expanded uncertainty of ±8.2% (99.7% level of confidence, k = 3). Bias was separately evaluated through a recovery study using standard reference material from a national metrology institute. The uncertainty estimate was verified through the use of proficiency test (PT) results. Assigned values for PT results and their associated uncertainties were calculated as robust means (x*) and standard deviations (s*) of participant values. Performance scores demonstrated that the uncertainty estimate was appropriate across the full range of PT concentrations (0.010-0.370 g/100 mL). The use of PT data as an empirical estimate of uncertainty was not examined. Until providers of blood ethanol PT samples include details on how an assigned value is obtained along with its uncertainty and traceability, the use of PT data should be restricted to the role of verification of uncertainty estimates.

  17. Validation of the Oregon Scientific BPU 330 for self-monitoring of blood pressure according to the International Protocol

    Directory of Open Access Journals (Sweden)

    Li Li

    2008-10-01

    Full Text Available Li Li1, XinYu Zhang1, ChunHong Yan1, QingXiang Liang21Biomedical Engineering Lab, Faculty of Information Engineering, ShenZhen University, ShenZhen, China; 2Bao An People’s Hospital, ShenZhen, ChinaObjective: Extensive marketing of devices for self-measurement of blood pressure has created a need for purchasers to be able to satisfy themselves that such devices have been evaluated according to agreed criteria. The Oregon Scientific BPU 330 blood pressure monitor is an electronic device for upper arm measurement. This study assessed the accuracy of the Oregon Scientific BPU 330 blood pressure monitor according to the International Protocol by the Working Group on Blood Pressure Monitoring of the European Society of Hypertension for validation of blood pressure measuring devices.Method: 52 participants over 30 years of age were studied in the validation. Nine blood pressure measurements were taken alternately with a mercury sphygmomanometer by two observers, and by the supervisor, using the BPU 330 device. A total of 33 participants were selected for the analysis. The validation was divided into two phases. Phase 1 included 15 participants. If the device passed phase 1, 18 more participants were included. The 99 pairs of measurements were compared according to the International Protocol. The device was given a pass/fail recommendation based on its accuracy compared with the mercury standard (within 5, 10, and 15 mmHg, as well as the number met in the ranges specified by the International Protocol.Results: The mean and standard deviation of the difference between the mean of the observers and the BPU 330 device were 1.7 ± 4.7 mmHg and 2.8 ± 3.9 mmHg for systolic blood pressure (SBP and diastolic blood pressure (DBP, respectively. In phase 1, the device passed with a total of 33, 43, and 44 SBP readings; 38, 44, and 45 DBP readings were within 5, 10, and 15 mmHg, respectively. In phase 2.1, 81, 95, and 96 for SBP, and 83, 95, and 98 for DBP

  18. Air Monitoring System in Elders' Apartment with QCM Type Gas Sensors

    Science.gov (United States)

    Kikuchi, Masashi; Ito, Tsukasa; Shiratori, Seimei

    The gas monitoring system for elders' apartment using QCM sensors was newly developed. The QCM sensors for sulfide gas and ammonia gas were used for this system. The system for bodily wastes was fabricated and applied to nursing care system in elders' apartment. This system is composed by the sensor unit, communication unit and data server. Care person can see whether the linen should be changed or not without seeing over each room. The QCM sensors have some problems such as the interference of humidity and temperature, therefore these influences were dissolved using humidity sensor and temperature sensor as feedback source. The sensors were placed in several points of elders' apartment for 2 weeks. This system can be used in elders' apartment successfully.

  19. Studies on automatic hot gas reader used in the countrywide personnel monitoring programme

    International Nuclear Information System (INIS)

    In India, ∼58,000 radiation workers are monitored using locally made CaSO4:Dy teflon embedded thermoluminescence dosemeter (TLD) badge system. The automatic hot gas readers developed locally are also used in TL measurements. The hot gas reader system has many advantages over the manual readers used previously and has completely replaced the manual reader system in all TLD personnel monitoring units in India. In the present study, the new reader system is studied and a theoretical attempt has been made to interpret the experimentally obtained results. The glow curves are generated theoretically and are also plotted experimentally. It has been found that the heat capacity of the heating gas, which is responsible for the transfer of heat, has a role in deciding the position of peak and is verified experimentally using different gas flow rates of nitrogen and argon as heating gases in the reader. The theoretical study may also be helpful in fitting the experimentally obtained glow curves and, therefore, the elimination of unwanted non-radiation-induced contributions, such as dark current, electronic spikes, light leakage and tribo-luminescence that generally distort the glow curve shape, can be achieved. (authors)

  20. Idaho National Laboratory (INL) Site Greenhouse Gas (GHG) Monitoring Plan - 40 CFR 98

    Energy Technology Data Exchange (ETDEWEB)

    Deborah L. Layton; Kimberly Frerichs

    2010-07-01

    The purpose of this Greenhouse Gas (GHG) Monitoring Plan is to meet the monitoring plan requirements of Title 40 of the Code of Federal Regulations Part 98.3(g)(5). This GHG Monitoring Plan identifies procedures and methodologies used at the Idaho National Laboratory Site (INL Site) to collect data used for GHG emissions calculations and reporting requirements from stationary combustion and other regulated sources in accordance with 40 CFR 98, Subparts A and other applicable subparts. INL Site Contractors determined subpart applicability through the use of a checklist (Appendix A). Each facility/contractor reviews operations to determine which subparts are applicable and the results are compiled to determine which subparts are applicable to the INL Site. This plan is applicable to the 40 CFR 98-regulated activities managed by the INL Site contractors: Idaho National Laboratory (INL), Idaho Cleanup Project (ICP), Advanced Mixed Waste Treatment Project (AMWTP), and Naval Reactors Facilities (NRF).

  1. Idaho National Laboratory (INL) Site Greenhouse Gas (GHG) Monitoring Plan - 40 CFR 98

    Energy Technology Data Exchange (ETDEWEB)

    Deborah L. Layton; Kimberly Frerichs

    2011-12-01

    The purpose of this Greenhouse Gas (GHG) Monitoring Plan is to meet the monitoring plan requirements of Title 40 of the Code of Federal Regulations Part 98.3(g)(5). This GHG Monitoring Plan identifies procedures and methodologies used at the Idaho National Laboratory Site (INL Site) to collect data used for GHG emissions calculations and reporting requirements from stationary combustion and other regulated sources in accordance with 40 CFR 98, Subparts A and other applicable subparts. INL Site Contractors determined subpart applicability through the use of a checklist (Appendix A). Each facility/contractor reviews operations to determine which subparts are applicable and the results are compiled to determine which subparts are applicable to the INL Site. This plan is applicable to the 40 CFR 98-regulated activities managed by the INL Site contractors: Idaho National Laboratory (INL), Idaho Cleanup Project (ICP), Advanced Mixed Waste Treatment Project (AMWTP), and Naval Reactors Facilities (NRF).

  2. Idaho National Laboratory (INL) Site Greenhouse Gas (GHG) Monitoring Plan - 40 CFR 98

    International Nuclear Information System (INIS)

    The purpose of this Greenhouse Gas (GHG) Monitoring Plan is to meet the monitoring plan requirements of Title 40 of the Code of Federal Regulations Part 98.3(g)(5). This GHG Monitoring Plan identifies procedures and methodologies used at the Idaho National Laboratory Site (INL Site) to collect data used for GHG emissions calculations and reporting requirements from stationary combustion and other regulated sources in accordance with 40 CFR 98, Subparts A and other applicable subparts. INL Site Contractors determined subpart applicability through the use of a checklist (Appendix A). Each facility/contractor reviews operations to determine which subparts are applicable and the results are compiled to determine which subparts are applicable to the INL Site. This plan is applicable to the 40 CFR 98-regulated activities managed by the INL Site contractors: Idaho National Laboratory (INL), Idaho Cleanup Project (ICP), Advanced Mixed Waste Treatment Project (AMWTP), and Naval Reactors Facilities (NRF).

  3. Enabling Technology for Monitoring & Predicting Gas Turbine Health & Performance in COAL IGCC Powerplants

    Energy Technology Data Exchange (ETDEWEB)

    Kenneth A. Yackly

    2004-09-30

    The ''Enabling & Information Technology To Increase RAM for Advanced Powerplants'' program, by DOE request, has been re-directed, de-scoped to two tasks, shortened to a 2-year period of performance, and refocused to develop, validate and accelerate the commercial use of enabling materials technologies and sensors for Coal IGCC powerplants. The new program has been re-titled as ''Enabling Technology for Monitoring & Predicting Gas Turbine Health & Performance in IGCC Powerplants'' to better match the new scope. This technical progress report summarizes the work accomplished in the reporting period April 1, 2004 to August 31, 2004 on the revised Re-Directed and De-Scoped program activity. The program Tasks are: Task 1--IGCC Environmental Impact on high Temperature Materials: This first materials task has been refocused to address Coal IGCC environmental impacts on high temperature materials use in gas turbines and remains in the program. This task will screen material performance and quantify the effects of high temperature erosion and corrosion of hot gas path materials in Coal IGCC applications. The materials of interest will include those in current service as well as advanced, high-performance alloys and coatings. Task 2--Material In-Service Health Monitoring: This second task develops and demonstrates new sensor technologies to determine the in-service health of advanced technology Coal IGCC powerplants, and remains in the program with a reduced scope. Its focus is now on only two critical sensor need areas for advanced Coal IGCC gas turbines: (1) Fuel Quality Sensor for detection of fuel impurities that could lead to rapid component degradation, and a Fuel Heating Value Sensor to rapidly determine the fuel heating value for more precise control of the gas turbine, and (2) Infra-Red Pyrometer to continuously measure the temperature of gas turbine buckets, nozzles, and combustor hardware.

  4. Influence of transcutaneous electrical stimulation on heterotopic ossification: an experimental study in Wistar rats.

    Science.gov (United States)

    Zotz, T G G; Paula, J B de

    2015-11-01

    Heterotopic ossification (HO) is a metaplastic biological process in which there is newly formed bone in soft tissues, resulting in joint mobility deficit and pain. Different treatment modalities have been tried to prevent HO development, but there is no consensus on a therapeutic approach. Since electrical stimulation is a widely used resource in physiotherapy practice to stimulate joint mobility, with analgesic and anti-inflammatory effects, its usefulness for HO treatment was investigated. We aimed to identify the influence of electrical stimulation on induced HO in Wistar rats. Thirty-six male rats (350-390 g) were used, and all animals were anesthetized for blood sampling before HO induction, to quantify the serum alkaline phosphatase. HO induction was performed by bone marrow implantation in both quadriceps of the animals, which were then divided into 3 groups: control (CG), transcutaneous electrical nerve stimulation (TENS) group (TG), and functional electrical stimulation (FES) group (FG) with 12 rats each. All animals were anesthetized and electrically stimulated twice per week, for 35 days from induction day. After this period, another blood sample was collected and quadriceps muscles were bilaterally removed for histological and calcium analysis and the rats were killed. Calcium levels in muscles showed significantly lower results when comparing TG and FG (P<0.001) and between TG and CG (P<0.001). Qualitative histological analyses confirmed 100% HO in FG and CG, while in TG the HO was detected in 54.5% of the animals. The effects of the muscle contractions caused by FES increased HO, while anti-inflammatory effects of TENS reduced HO. PMID:26292223

  5. Penile rehabilitation with a vacuum erectile device in an animal model is related to an antihypoxic mechanism: blood gas evidence.

    Science.gov (United States)

    Lin, Hao-Cheng; Yang, Wen-Li; Zhang, Jun-Lan; Dai, Yu-Tian; Wang, Run

    2013-05-01

    Our previous study showed that vacuum erectile device (VED) therapy has improved erectile function in rats with bilateral cavernous nerve crush (BCNC) injuries. This study was designed to explore the mechanism of VED in penile rehabilitation by analyzing cavernous oxygen saturation (SO2) and to examine the effect of VED therapy on preventing penile shrinkage after BCNC. Thirty adult Sprague-Dawley rats were randomly assigned into three groups: group 1, sham surgery; group 2, BCNC; and group 3, BCNC+VED. Penile length and diameter were measured on a weekly basis. After 4 weeks of therapy, the penile blood was extracted by three methods for blood gas analysis (BGA): method 1, cavernous blood was aspirated at the flaccid state; method 2, cavernous blood was aspirated at the traction state; and method 3, cavernous blood was aspirated immediately after applying VED. SO2 values were tested by the blood gas analyzer. The results showed that VED therapy is effective in preventing penile shrinkage induced by BCNC (Penile shortening: BCNC group 1.9±1.1 mm; VED group 0.3±1.0 mm; PVED group 0.04±0.14 mm; PVED application (88.25%±4.94%) compared to the flaccid (76.53%±4.16%) or traction groups (78.93%±2.56%) (PVED application were 62% arterial and 38% venous blood. These findings suggest that VED therapy can effectively preserve penile size in rats with BCNC injury. The beneficial effect of VED therapy is related to antihypoxia by increasing cavernous blood SO2.

  6. Penile rehabilitation with a vacuum erectile device in an animal model is related to an antihypoxic mechanism: blood gas evidence.

    Science.gov (United States)

    Lin, Hao-Cheng; Yang, Wen-Li; Zhang, Jun-Lan; Dai, Yu-Tian; Wang, Run

    2013-05-01

    Our previous study showed that vacuum erectile device (VED) therapy has improved erectile function in rats with bilateral cavernous nerve crush (BCNC) injuries. This study was designed to explore the mechanism of VED in penile rehabilitation by analyzing cavernous oxygen saturation (SO2) and to examine the effect of VED therapy on preventing penile shrinkage after BCNC. Thirty adult Sprague-Dawley rats were randomly assigned into three groups: group 1, sham surgery; group 2, BCNC; and group 3, BCNC+VED. Penile length and diameter were measured on a weekly basis. After 4 weeks of therapy, the penile blood was extracted by three methods for blood gas analysis (BGA): method 1, cavernous blood was aspirated at the flaccid state; method 2, cavernous blood was aspirated at the traction state; and method 3, cavernous blood was aspirated immediately after applying VED. SO2 values were tested by the blood gas analyzer. The results showed that VED therapy is effective in preventing penile shrinkage induced by BCNC (Penile shortening: BCNC group 1.9±1.1 mm; VED group 0.3±1.0 mm; PVED group 0.04±0.14 mm; PVED application (88.25%±4.94%) compared to the flaccid (76.53%±4.16%) or traction groups (78.93%±2.56%) (PVED application were 62% arterial and 38% venous blood. These findings suggest that VED therapy can effectively preserve penile size in rats with BCNC injury. The beneficial effect of VED therapy is related to antihypoxia by increasing cavernous blood SO2. PMID:23564044

  7. Simultaneous determination of polycyclic musks in blood and urine by solid supported liquid-liquid extraction and gas chromatography-tandem mass spectrometry.

    Science.gov (United States)

    Liu, Hongtao; Huang, Liping; Chen, Yuxin; Guo, Liman; Li, Limin; Zhou, Haiyun; Luan, Tiangang

    2015-06-15

    A rapid, precise and accurate method for the simultaneous determination of 5 polycyclic musks (PCMs) in biological fluids was developed by solid supported liquid-liquid extraction (SLE) coupled with gas chromatography-tandem mass spectrometry (GC-MS/MS). All parameters influencing SLE-GC-MS performance, including electron energy of electron-impact ionization source, collision energy for tandem mass spectrometer when operated in selected-reaction monitoring (SRM) mode, type and volume of elution reagent, nitrogen evaporation time, pH and salinity of sample have been carefully optimized. Eight milliliter of n-hexane was finally chosen as elution reagent. Blood and urine sample could be loaded into SLE cartridge without adjusting pH and salinity. Deuterated tonalide (AHTN-d3) was chosen as internal standard. The correlation coefficient (r(2)) of the calibration curves of target compounds ranged from 0.9996 to 0.9998. The dynamic range spanned over two orders of magnitude. The limit of detection (LOD) of target compounds in blood and urine ranged from 0.008 to 0.105μgL(-1) and 0.005 to 0.075μgL(-1), respectively. The developed procedure was successfully applied to the analysis of PCMs in human blood and urine obtaining satisfying recoveries on low, medium and high levels. The method was compared with SLE-GC-MS and shown one to two orders of magnitude improvement in sensitivity.

  8. One year continuous soil gas monitoring above an EGR test site

    Science.gov (United States)

    Furche, Markus; Schlömer, Stefan; Faber, Eckhard; Dumke, Ingolf

    2010-05-01

    Setup and first results of an ongoing research activity are presented, which is funded by the German Geotechnologien program within in the joint project CLEAN (CO2 Large Scale Enhanced Gas Recovery in the Altmark Natural Gas Field). The task is to establish several soil gas monitoring stations above a partly exhausted gas field in the Altmark which will be used for an enhanced gas recovery (EGR) test by injecting CO2 into the reservoir. The aim is to optimize the monitoring technique including automatic data transfer and data exploitation and to understand mechanisms of natural variations of soil gas concentrations in the specific area. Furthermore the suitability of these measurements as a contribution to leakage detection shall be evaluated. A network of 13 gauging stations for the measurement of CO2 is working continuously for about one year. They are spread over an area of 8 x 3 km and are situated in direct vicinity of existing deep boreholes as the most likely locations for possible leakage. In addition one station is placed far outside the gasfield as a reference point. The technique applied to measure soil gas concentrations uses a gas stream circulating in a tube going down a shallow borehole where the circulating gas is in contact with the soil gas phase via a gas permeable membrane. Above surface, moisture is removed from the gas stream before it reaches several gas sensors for CO2. Besides these, several other parameters are determined as well, e.g. soil moisture and soil temperature, water level, gas flow and gas moisture. In addition a meteorological station gives information about precipitation, air humidity, temperature and pressure, global radiation, wind direction and velocity in the area. Data are continuously collected by dataloggers at each station (5 minutes interval), transferred via GSM routers to the BGR server in Hannover and are stored in a specially designed database. The database does not only contain the measurements but also

  9. [Effects of procaterol on arterial blood gas and pulmonary function in asthmatic children].

    Science.gov (United States)

    Obata, T; Masaki, T; Iikura, Y

    1990-12-01

    We examined the changes in arterial blood gas, FEV1 and V50 after the inhalation of procaterol on 19 occasions in 16 asthmatic children. The initial value of PaO2 had statistically significant correlation with the initial values of %FEV1 and %V50. PaO2 fell down in 11 out of 19 (57.9%) and remarkably decreased more than 5 mmHg in 6 out of 19 (31.6%). The fall in PaO2 was most significant at 5 min after the inhalation in almost subjects. The changes in PaO2 after the inhalation had good correlation with the initial value of %FEV1 and %V50. The initial values were quite lower in the patients with decreased PaO2 more than 5 mmHg than those with increased PaO2. Severe patients showed statistically low values of the initial PaO2 and %V50 and showed a fall in PaO2 after inhalation compared with moderate patients. An increase in A-aDO2 elicited that deteriorations of V/Q ratio caused a decrease in PaO2 after inhalation. There were no significant changes in heart rates and no complaints of nausea, headache or tremor.

  10. Monitoring induced seismicity from underground gas storage: first steps in Italy

    Science.gov (United States)

    Mucciarelli, Marco; Priolo, Enrico

    2013-04-01

    The supply of natural gas and its storage are focal points of the Italian politics of energy production and will have increasing importance in the coming years. About a dozen reservoirs are currently in use and fifteen are in development or awaiting approval. Some of these are found in the vicinity of geological structures that are seismically active. The assessment of seismic hazard (both for natural background and induced seismicity) for a geological gas storage facility has a number of unconventional aspects that must be recognized and traced in a clear, ordered way and using guidelines and rules that leave less room as possible for interpretation by the individual applicant / verification body. Similarly, for control and monitoring there are not clearly defined procedures or standard instrumentation, let alone tools for analysing and processing data. Finally, governmental organizations in charge of permission grants and operative control tend to have appropriate scientific knowledge only in certain areas and not in others (e.g. the seismic one), and the establishment of an independent multidisciplinary inspection body appears desirable. The project StoHaz (https://sites.google.com/site/s2stohaz/home) aims to initiate a series of actions to overcome these deficiencies and allow to define procedures and standards for the seismic hazard assessment and control of the activities of natural gas storage in underground reservoirs. OGS will take advantage of the experience gained with the design, installation and maintenance of the seismic network monitoring the Collalto reservoir, at the moment the only example in Italy of a public research institution monitoring independently the activities of a private gas storage company.

  11. Results of a Long-Term Demonstration of an Optical Multi-Gas Monitor on ISS

    Science.gov (United States)

    Mudgett, Paul; Pilgrim, Jeffrey S.

    2015-01-01

    Previously at SAMAP we reported on the development of tunable diode laser spectroscopy (TDLS) based instruments for measuring small gas molecules in real time. TDLS technology has matured rapidly over the last 5 years as a result of advances in low power diode lasers as well as better detection schemes. In collaboration with two small businesses Vista Photonics, Inc. and Nanoracks LLC, NASA developed a 4 gas TDLS based monitor for an experimental demonstration of the technology on the International Space Station (ISS). Vista invented and constructed the core TDLS sensor. Nanoracks designed and built the enclosure, and certified the integrated monitor as a payload. The device, which measures oxygen, carbon dioxide, ammonia and water vapor, is called the Multi-Gas Monitor (MGM). MGM measures the 4 gases every few seconds and records a 30 second moving average of the concentrations. The relatively small unit draws only 2.5W. MGM was calibrated at NASA-Johnson Space Center in July 2013 and launched to ISS on a Soyuz vehicle in November 2013. Installation and activation of MGM occurred in February 2014, and the unit has been operating nearly continuously ever since in the Japanese Experiment Module. Data is downlinked from ISS about once per week. Oxygen and carbon dioxide data is compared with that from the central Major Constituents Analyzer. Water vapor data is compared with dew point measurements made by sensors in the Columbus module. The ammonia channel was tested by the crew using a commercial ammonia inhalant. MGM is remarkably stable to date. Results of 18 months of operation are presented and future applications including combustion product monitoring are discussed.

  12. Consensus report: the current role of self-monitoring of blood glucose in non-insulin-treated type 2 diabetes.

    Science.gov (United States)

    Klonoff, David C; Blonde, Lawrence; Cembrowski, George; Chacra, Antonio Roberto; Charpentier, Guillaume; Colagiuri, Stephen; Dailey, George; Gabbay, Robert A; Heinemann, Lutz; Kerr, David; Nicolucci, Antonio; Polonsky, William; Schnell, Oliver; Vigersky, Robert; Yale, Jean-François

    2011-11-01

    The Coalition for Clinical Research--Self-Monitoring of Blood Glucose Scientific Board convened a meeting in San Francisco, CA, July 20-21, 2011, to discuss the current practice of self-monitoring of blood glucose (SMBG) in non-insulin-treated (NIT) type 2 diabetes mellitus (T2DM). Twelve physician panel members from academia, practice, and government attended this meeting. These experts came from the United States, Brazil, Canada, France, Germany, Italy, and the United Kingdom. In addition, three consultants from Australia, Germany, and the United States contributed to the group's final report. This coalition was organized by Diabetes Technology Society. Self-monitoring of blood glucose was studied from eight perspectives related to patients with NIT T2DM: (1) epidemiological studies; (2) randomized controlled trials (RCT)s and meta-analyses; (3) targets, timing, and frequency of SMBG use; (4) incidence and role of SMBG in preventing hypoglycemia with single-drug regimens and combination regimens consisting of antihyperglycemic agents other than secretagogues and insulin; (5) comparison of SMBG with continuous glucose monitoring; (6) technological capabilities and limitations of SMBG; (7) barriers to appropriate use of SMBG; and (8) methods and end points for appropriate future clinical trials. The panel emphasized recent studies, which reflect the current approach for applying this intervention. Among the participants there was consensus that: SMBG is an established practice for patients with NIT T2DM, and to be most effective, it should be performed in a structured format where information obtained from this measurement is used to guide treatment; New, high-quality efficacy data from RCTs have demonstrated efficacy of SMBG in NIT T2DM in trials reported since 2008; Both patients and health care professionals require education on how to respond to the data for SMBG to be effective; and Additional well-defined studies are needed to assess the benefits and costs of

  13. Barometric gas transport along faults and its application to nuclear test-ban monitoring

    International Nuclear Information System (INIS)

    Underground nuclear explosions produce a unique but evanescent set of radionuclide gases that potentially can be used in the context of an on-site, test-ban monitoring program to differentiate them from other detected events such as earthquakes or mining activity. In Part I of this report we describe an experiment to evaluate the upward transport of gases from an underground explosion using two gas tracers with very different diffusivities that were released in a 400-m-deep, chemical explosive detonation. The less diffusive (more massive) tracer was detected on a nearby geologic fault 50 days following the detonation while the more diffusive tracer was detected 375 days after release. Computer simulations indicate that the arrival time and the chromatographic behavior of transport are characteristic of barometrically induced flow in a fractured, porous matrix regime. For a hypothetical 1-kiloton fission explosion subject to the same weather and gas transport conditions of the chemical explosion, simulations predict the detectability of argon-37 after 80 days in spite of depletion by radioactive decay. Largely because of the earlier arrival of xenon-133, owing to its lower binary gas diffusivity, the exceedingly short lived isotope should also be detectable - arriving about 30 days earlier than argon. In Part II we consider that our prediction of the detectability of argon and xenon is based upon the small volume (0.00001 m3) sampling technique of the NPE tracer-gas sampling study while actual sampling for radionuclides would involve drawing much larger volume (possibly 0.1-1 m3) gas samples from the near-surface. Extraction of such a large volume of gas from 1-5 meter depths in the soil raises the possibility of significant atmospheric infiltration, leading to substantial dilution of the extracted gas sample. However, an infiltration experiment suggests that significant dilution would not, in fact, occur at the most prolific sampling stations of the earlier gas

  14. Use of UAVs for greenhouse gas monitoring at hotspot emissions zones

    Science.gov (United States)

    Pitt, J. R.; Allen, G.; Mead, M. I.; Hollingsworth, P.; Kabbabe, K.; Roberts, G.; Shallcross, D. E.

    2015-12-01

    Measuring greenhouse gas emissions from individual localised sources, or "hotspots", is important for both compliance monitoring and validating the techniques used to compile national emission inventories. Frequently ground based techniques are used, such as flux chamber measurements, which suffer from issues regarding sample representativeness, and tracer release methods, which for area sources rely heavily on release site configuration. Obtaining vertically resolved data can enable the use of a mass balance method to calculate greenhouse gas fluxes. This has been achieved using remote sensing techniques, but this usually requires the deployment of expensive, bulky instrumentation. Here we evaluate the suitability of using UAVs, in conjunction with emerging miniaturised sensor technology, as a highly manoeuvrable, low cost alternative for measuring hotspot greenhouse gas emissions. We describe a case study performed at a UK landfill site, where greenhouse gas measurements made on board a fixed wing UAV were used to estimate the bulk CH4 emission rate. Details of the mass balance technique employed, along with the key uncertainties associated with it, are discussed. This work is part of an ongoing study at the University of Manchester into the application of UAVs in atmospheric research, with the rapid advancement in miniaturised sensor technology providing new opportunities for integrating trace gas measurement with existing lightweight UAVs.

  15. On-Line Fission Gas Release Monitoring System in the High Flux Reactor Petten

    International Nuclear Information System (INIS)

    For HTR fuel irradiation tests in the HFR Petten a specific installation was designed and installed dubbed the 'Sweep Loop Facility' (SLF). The SLF is tasked with three functions, namely temperature control by gas mixture technique, surveillance of safety parameters (temperature, pressure, radioactivity etc.) and analysis of fission gas release for three individual capsules in two separate experimental rigs. The SLF enables continuous and independent surveillance of all gas circuits. The release of volatile fission products (FP) from the in-pile experiments is monitored by continuous gas purging. The fractional release of these FP, defined as the ratio between release rate of a gaseous fission isotope (measured) to its instantaneous birth rate (calculated), is a licensing-relevant test for HTR fuel. The developed gamma spectrometry station allows for higher measurement frequencies, thus enabling follow-up of rapid and massive release transients. The designed stand-alone system was tested and fully used through the final irradiation period of the HFR-EU1 experiment which was terminated on 18 February 2010. Its robustness allowed the set up to be used as extra safety instrumentation. This paper describes the gas activity measurement technique based on HPGe gamma spectrometry and illustrates how qualitative and quantitative analysis of volatile FP can be performed on-line. (authors)

  16. Tolerability of the Oscar 2 ambulatory blood pressure monitor among research participants: a cross-sectional repeated measures study

    Directory of Open Access Journals (Sweden)

    Hinderliter Alan L

    2011-04-01

    Full Text Available Abstract Background Ambulatory blood pressure monitoring (ABPM is increasingly used to measure blood pressure (BP in research studies. We examined ease of use, comfort, degree of disturbance, reported adverse effects, factors associated with poor tolerability, and association of poor tolerability with data acquisition of 24-hour ABPM using the Oscar 2 monitor in the research setting. Methods Sixty adults participating in a research study of people with a history of borderline clinic BP reported on their experience with ABPM on two occasions one week apart. Poor tolerability was operationalized as an overall score at or above the 75th percentile using responses to questions adapted from a previously developed questionnaire. In addition to descriptive statistics (means for responses to Likert-scaled "0 to 10" questions and proportions for Yes/No questions, we examined reproducibility of poor tolerability as well as associations with poor tolerability and whether poor tolerability was associated with removal of the monitor or inadequate number of BP measurements. Results The mean ambulatory BP of participants by an initial ABPM session was 148/87 mm Hg. After wearing the monitor the first time, the degree to which the monitor was felt to be cumbersome ranged from a mean of 3.0 to 3.8, depending on whether at work, home, driving, or other times. The most bother was interference with normal sleeping pattern (mean 4.2. Wearers found the monitor straightforward to use (mean 7.5. Nearly 67% reported that the monitor woke them after falling asleep, and 8.6% removed it at some point during the night. Reported adverse effects included pain (32%, skin irritation (37%, and bruising (7%. Those categorized as having poor tolerability (kappa = 0.5 between sessions, p = 0.0003 were more likely to report being in fair/poor health (75% vs 22%, p = 0.01 and have elevated 24-hour BP average (systolic: 28% vs 17%, p = 0.56; diastolic: 30% vs 17%, p = 0.37. They were

  17. Industrial SO2 emission monitoring through a portable multichannel gas analyzer with an optimized retrieval algorithm

    Science.gov (United States)

    Sun, Youwen; Liu, Cheng; Xie, Pinhua; Hartl, Andreas; Chan, Kalok; Tian, Yuan; Wang, Wei; Qin, Min; Liu, Jianguo; Liu, Wenqing

    2016-03-01

    SO2 variability over a large concentration range and interferences from other gases have been major limitations in industrial SO2 emission monitoring. This study demonstrates accurate industrial SO2 emission monitoring through a portable multichannel gas analyzer with an optimized retrieval algorithm. The proposed analyzer features a large dynamic measurement range and correction of interferences from other coexisting infrared absorbers such as NO, CO, CO2, NO2, CH4, HC, N2O, and H2O. The multichannel gas analyzer measures 11 different wavelength channels simultaneously to correct several major problems of an infrared gas analyzer including system drift, conflict of sensitivity, interferences among different infrared absorbers, and limitation of measurement range. The optimized algorithm uses a third polynomial instead of a constant factor to quantify gas-to-gas interference. Measurement results show good performance in the linear and nonlinear ranges, thereby solving the problem that the conventional interference correction is restricted by the linearity of the intended and interfering channels. The results imply that the measurement range of the developed multichannel analyzer can be extended to the nonlinear absorption region. The measurement range and accuracy are evaluated through experimental laboratory calibration. Excellent agreement was achieved, with a Pearson correlation coefficient (r2) of 0.99977 with a measurement range from approximately 5 to 10 000 ppmv and a measurement error of less than 2 %. The instrument was also deployed for field measurement. Emissions from three different factories were measured. The emissions of these factories have been characterized by different coexisting infrared absorbers, covering a wide range of concentration levels. We compared our measurements with commercial SO2 analyzers. Overall, good agreement was achieved.

  18. Tissue-Informative Mechanism for Wearable Non-invasive Continuous Blood Pressure Monitoring

    OpenAIRE

    Woo, Sung Hun; Choi, Yun Young; Kim, Dae Jung; Bien, Franklin; Kim, Jae Joon

    2014-01-01

    Accurate continuous direct measurement of the blood pressure is currently available thru direct invasive methods via intravascular needles, and is mostly limited to use during surgical procedures or in the intensive care unit (ICU). Non-invasive methods that are mostly based on auscultation or cuff oscillometric principles do provide relatively accurate measurement of blood pressure. However, they mostly involve physical inconveniences such as pressure or stress on the human body. Here, we in...

  19. Bird interactions with offshore oil and gas platforms: review of impacts and monitoring techniques.

    Science.gov (United States)

    Ronconi, Robert A; Allard, Karel A; Taylor, Philip D

    2015-01-01

    Thousands of oil and gas platforms are currently operating in offshore waters globally, and this industry is expected to expand in coming decades. Although the potential environmental impacts of offshore oil and gas activities are widely recognized, there is limited understanding of their impacts on migratory and resident birds. A literature review identified 24 studies and reports of bird-platform interactions, most being qualitative and half having been peer-reviewed. The most frequently observed effect, for seabirds and landbirds, is attraction and sometimes collisions associated with lights and flares; episodic events have caused the deaths of hundreds or even thousands of birds. Though typically unpredictable, anecdotally, it is known that poor weather, such as fog, precipitation and low cloud cover, can exacerbate the effect of nocturnal attraction to lights, especially when coincidental with bird migrations. Other effects include provision of foraging and roosting opportunities, increased exposure to oil and hazardous environments, increased exposure to predators, or repulsion from feeding sites. Current approaches to monitoring birds at offshore platforms have focused on observer-based methods which can offer species-level bird identification, quantify seasonal patterns of relative abundance and distribution, and document avian mortality events and underlying factors. Observer-based monitoring is time-intensive, limited in spatial and temporal coverage, and suffers without clear protocols and when not conducted by trained, independent observers. These difficulties are exacerbated because deleterious bird-platform interaction is episodic and likely requires the coincidence of multiple factors (e.g., darkness, cloud, fog, rain conditions, occurrence of birds in vicinity). Collectively, these considerations suggest a need to implement supplemental systems for monitoring bird activities around offshore platforms. Instrument-based approaches, such as radar

  20. Wireless Sensor Network in Niger Delta Oil and Gas Field Monitoring: The Security Challenges and Countermeasures

    Directory of Open Access Journals (Sweden)

    Fidelis C. Obodoeze

    2012-12-01

    Full Text Available The IEEE 802.15.4 specification has enabled low-power, low-cost and smart wireless sensor networks(WSNs capable of robust and reliable multi-hop communications. By January 2005, an International Oiland Gas Company (IOC, Shell Petroleum Development Company (SPDC, became the firstmultinational Oil and Gas Company operating in the Nigeria Niger Delta region to switch from wired towireless sensor technology eliminating the need for cables thereby allowing data collection in remote,swampy areas and enabling new applications. However, there are concerns related to the use of thesesmart wireless sensor networks such as reliability, standardization, energy consumption and generaloperational, data and physical security issues especially in the monitoring of mission-critical oil and gasinstallations and infrastructure such as pipelines, oil wells, oil rigs and flow stations in a regioncharacterized by rampant vandalisation and sabotage of oil pipelines and other oil installations bymilitants and oil thieves. High cases of vandalisation of oil and gas pipelines and other oil installationswere identified even when there is evidence of wireless sensor deployment. This paper introducedpractical deployment architectures and mechanisms that can secure oil facilities and the wireless sensorsfrom being physically attacked so that they can successfully monitor and report incidences of pipeline andequipment vandalisation easily and on time while at the same time maintain data security of the WSN.

  1. Physics Simulation Software for Autonomous Propellant Loading and Gas House Autonomous System Monitoring

    Science.gov (United States)

    Regalado Reyes, Bjorn Constant

    2015-01-01

    1. Kennedy Space Center (KSC) is developing a mobile launching system with autonomous propellant loading capabilities for liquid-fueled rockets. An autonomous system will be responsible for monitoring and controlling the storage, loading and transferring of cryogenic propellants. The Physics Simulation Software will reproduce the sensor data seen during the delivery of cryogenic fluids including valve positions, pressures, temperatures and flow rates. The simulator will provide insight into the functionality of the propellant systems and demonstrate the effects of potential faults. This will provide verification of the communications protocols and the autonomous system control. 2. The High Pressure Gas Facility (HPGF) stores and distributes hydrogen, nitrogen, helium and high pressure air. The hydrogen and nitrogen are stored in cryogenic liquid state. The cryogenic fluids pose several hazards to operators and the storage and transfer equipment. Constant monitoring of pressures, temperatures and flow rates are required in order to maintain the safety of personnel and equipment during the handling and storage of these commodities. The Gas House Autonomous System Monitoring software will be responsible for constantly observing and recording sensor data, identifying and predicting faults and relaying hazard and operational information to the operators.

  2. Reliable long-term continuous blood glucose monitoring for patients in critical care using microdialysis and infrared spectrometry

    Science.gov (United States)

    Heise, H. Michael; Damm, Uwe; Kondepati, Venkata R.

    2006-02-01

    For clinical research, in-vivo blood glucose monitoring is an ongoing important topic to improve glycemic control in patients with non-adequate blood glucose regulation. Critically ill patients received much interest, since the intensive insulin therapy treatment, as established for diabetics, reduces mortality significantly. Despite the existence of commercially available, mainly amperometric biosensors, continued interest is in infrared spectroscopic techniques for reagent-free glucose monitoring. For stable long-term operation, avoiding also sensor recalibration, a bed-side device coupled to a micro-dialysis probe was developed for quasi-continuous glucose monitoring. Multivariate calibration is required for glucose concentration prediction due to the complex composition of dialysates from interstitial body fluid. Measurements were carried out with different test persons, each experiment lasting for more than 8 hours. Owing to low dialysis recovery rates, glucose concentrations in the dialysates were between 0.83 and 4.44 mM. Standard errors of prediction (SEP) obtained with Partial Least Squares (PLS) calibration and different cross-validation strategies were mainly between 0.13 and 0.18 mM based on either full interval data or specially selected spectral variables.

  3. Cost-benefit analysis of home blood pressure monitoring in hypertension diagnosis and treatment: an insurer perspective.

    Science.gov (United States)

    Arrieta, Alejandro; Woods, John R; Qiao, Nan; Jay, Stephen J

    2014-10-01

    Home blood pressure (BP) monitoring has been shown to be more effective than clinic BP monitoring for diagnosing and treating hypertension. However, reimbursement of home BP monitoring is uncommon in the United States because of a lack of evidence that it is cost beneficial for insurers. We develop a decision-analytic model, which we use to conduct a cost-benefit analysis from the perspective of the insurer. Model inputs are derived from the 2008 to 2011 claims data of a private health insurer in the United States, from 2009 to 2010 National Health and the Nutrition Examination Survey data, and from published meta-analyses. The model simulates the transitions among health states from initial physician visit to hypertension diagnosis, to treatment, to hypertension-related cardiovascular diseases, and patient death or resignation from the plan. We use the model to estimate cost-benefit ratios and both short- and long-run return on investment for home BP monitoring compared with clinic BP monitoring. Our results suggest that reimbursement of home BP monitoring is cost beneficial from an insurer's perspective for diagnosing and treating hypertension. Depending on the insurance plan and age group categories considered, estimated net savings associated with the use of home BP monitoring range from $33 to $166 per member in the first year and from $415 to $1364 in the long run (10 years). Return on investment ranges from $0.85 to $3.75 per dollar invested in the first year and from $7.50 to $19.34 per dollar invested in the long run.

  4. The monitoring system of the pipeline safety of the coal mine gas drainage based on the optic fiber sensing technology

    Science.gov (United States)

    Li, Yan-fang; Wei, Yubin; Zhang, Tingting; Zhao, Yanjie; Lv, Lei; Liu, Tong-yu

    2013-09-01

    Based on the technology of the spectrum absorption and the FBG, the monitoring system realize on line detection of the concentration of methane and oxygen, the temperature and the pressure of the gas in pipeline, and in order to improve the accuracy of the gas detection, we induce the compensation to the gas concentration using the data of the temperature and the pressure. In order to have a effective utilization of the methane in the coal mine gas drainage system, we have to have a accurate measurements of the concentration, the temperature and the pressure of the gas in pipeline. At the same time the dynamic monitoring of the concentration of Oxygen is a sign of the leakage of the pump. This paper gave some data detected in the field of the coal mine gas drainage system.

  5. Management of patients with uncontrolled arterial hypertension – the role of electronic compliance monitoring, 24-h ambulatory blood pressure monitoring and Candesartan/HCT

    Directory of Open Access Journals (Sweden)

    Tousset Eric

    2006-08-01

    Full Text Available Abstract Background Incomplete drug regimen compliance (DRC and white-coat hypertension are two of several possible causes of uncontrolled hypertension. Therefore the aim of the present study was to compare DRC in hypertensives treated with combination therapy whose blood pressures (BP were controlled vers. uncontrolled after 4 weeks of self-monitored BP measurement. To observe the consequences in uncontrolled patients of switching one drug of the combination therapy to candesartan/HCTZ (16 mg/12.5 mg with and without a compliance intervention program. Methods Self-and ambulatory-monitoring of BP were done with upper arm oscillometric devices. Patients' dosing histories were compiled electronically (MEMS(c, AARDEX. Patients with office blood pressure (OBP >140/90 mmHg despite combination therapy were begun on MEMS monitoring and self BP measurement for 4 weeks of run-in. Of 62 such patients, 18 (29% patients were normotensive according to self BP measurement and ambulatory BP measurement at 4 weeks (Group A; in the remaining 44 still uncontrolled patients, candesartan/HCTZ was substituted for one of the combination therapy drugs, with half these patients receiving passive compliance monitoring (B and half a DRC intervention program (C. All groups were then followed for 8 weeks. Results DRC before week 4 was significantly higher in A than in the uncontrolled patients (B&C. DRC was stable during run-in A, but declined in B and C. DRC after week 4 was not different in the three groups and stayed constant over time. DRC during weekends was lower than during weekdays in all groups. In group A no significant change in blood pressure was observed with all three methods of BP measurements. In groups B and C significant reductions of systolic and diastolic BP were observed for ABPM and SBPM. After the change to candesartan/HCTZ in B&C ambulatory 24-h-BP (ABPM was normalized in 39% of patients. Conclusion Normalization of BP was associated with superior

  6. Roles of preoperative arterial blood gas tests in the surgical treatment of scoliosis with moderate or severe pulmonary dysfunction

    Institute of Scientific and Technical Information of China (English)

    LIU Jia-ming; SHEN Jian-xiong; ZHANG Jian-guo; ZHAO Hong; LI Shu-gang; ZHAO Yu; QIU Giu-xing

    2012-01-01

    Background It has been stated that preoperative pulmonary function tests are essential to assess the surgical risk in patients with scoliosis.Arterial blood gas tests have also been used to evaluate pulmonary function before scoliotic surgery.However,few studies have been reported.The aim of this study was to investigate the roles of preoperative arterial blood gas tests in the surgical treatment of scoliosis with moderate or severe pulmonary dysfunction.Methods This study involved scoliotic patients with moderate or severe pulmonary dysfunction (forced vital capacity <60%) who underwent surgical treatment between January 2002 and April 2010.A total of 73 scoliotic patients (23 males and 50 females) with moderate or severe pulmonary dysfunction were included.The average age of the patients was 16.53 years (ranged 10-44).The demographic distribution,medical records,and radiographs of all patients were collected.All patients received arterial blood gas tests and pulmonary function tests before surgery.The arterial blood gas tests included five parameters:partial pressure of arterial oxygen,partial pressure of arterial carbon dioxide,alveolar-arterial oxygen tension gradient,pH,and standard bases excess.The pulmonary function tests included three parameters:forced expiratory volume in 1 second ratio,forced vital capacity ratio,and peak expiratory flow ratio.All five parameters of the arterial blood gas tests were compared between the two groups with or without postoperative pulmonary complications by variance analysis.Similarly,all three parameters of the pulmonary function tests were compared.Results The average coronal Cobb angle before surgery was 97.42° (range,50°-180°).A total of 15 (20.5%) patients had postoperative pulmonary complications,including hypoxemia in 5 cases (33.3%),increased requirement for postoperative ventilatory support in 4 (26.7%),pneumonia in 2 (13.3%),atelectasis in 2 (13.3%),pneumothorax in 1 (6.7%),and hydrothorax in 1

  7. Full-field speckle correlation technique as applied to blood flow monitoring

    Science.gov (United States)

    Vilensky, M. A.; Agafonov, D. N.; Timoshina, P. A.; Shipovskaya, O. V.; Zimnyakov, D. A.; Tuchin, V. V.; Novikov, P. A.

    2011-03-01

    The results of experimental study of monitoring the microcirculation in tissue superficial layers of the internal organs at gastro-duodenal hemorrhage with the use of laser speckles contrast analysis technique are presented. The microcirculation monitoring was provided in the course of the laparotomy of rat abdominal cavity in the real time. Microscopic hemodynamics was analyzed for small intestine and stomach under different conditions (normal state, provoked ischemia, administration of vasodilative agents such as papaverine, lidocaine). The prospects and problems of internal monitoring of micro-vascular flow in clinical conditions are discussed.

  8. Potential of optical microangiography to monitor cerebral blood perfusion and vascular plasticity following traumatic brain injury in mice in vivo

    Science.gov (United States)

    Jia, Yali; Alkayed, Nabil; Wang, Ruikang K.

    2009-07-01

    Optical microanglography (OMAG) is a recently developed imaging modality capable of volumetric imaging of dynamic blood perfusion, down to capillary level resolution, with an imaging depth up to 2.00 mm beneath the tissue surface. We report the use of OMAG to monitor the cerebral blood flow (CBF) over the cortex of mouse brain upon traumatic brain injury (TBI), with the cranium left intact, for a period of two weeks on the same animal. We show the ability of OMAG to repeatedly image 3-D cerebral vasculatures during pre- and post-traumatic phases, and to visualize the changes of regulated CBF and the vascular plasticity after TBI. The results indicate the potential of OMAG to explore the mechanism involved in the rehabilitation of TBI.

  9. Transcutaneous electrical nerve stimulation therapy in reduction of orofacial pain

    Directory of Open Access Journals (Sweden)

    Đorđević Igor

    2014-01-01

    Full Text Available Introduction. Patients with craniomandibular disorders suffer from hypertonic, fatigued and painful masticatory muscles. This condition can lead to limitation of mandibular jaw movements. All of these symptoms and signs are included in myofascial pain dysfunction syndrome. Transcutaneous electrical nerve stimulation (TENS has been used for treatment of these patients. Objective. The aim of this study was to assess the effect of TENS therapy on chronic pain reduction in patients with the muscular dysfunction symptom. Methods. In order to evaluate the effect of TENS therapy before and after the treatment, Craniomandibular Index (Helkimo was used. Pain intensity was measured by VAS. Patients had TENS treatment over two-week period. BURST TENS modality was used. Current intensity was individually adjusted. Results. Two patients did not respond to TENS therapy. Complete pain reduction was recorded in 8 patients, while pain reduction was not significantly different after TENS therapy in 10 patients. Conclusion. TENS therapy was confirmed as therapeutic procedure in orofacial muscle relaxation and pain reduction.

  10. Improvement of physical performance by transcutaneous nerve stimulation in athletes.

    Science.gov (United States)

    Kaada, B

    1984-01-01

    The present pilot study tested the exercise response to transcutaneous nerve stimulation (TNS) of 21 volunteers, who were well-trained competitive athletes. In 62 experiments (n) they received low-frequency TNS (2 Hz) for 30-45 min prior to either a road or track race, swimming race, bicycle ergometer exercise, isometric muscular endurance test, or dynamometer hand grip test. Improvement in performance compared with a corresponding number of control tests, without TNS or with placebo stimulation in the same subjects, was almost regularly observed in running, swimming and ergometer cycling, although with great individual variations. The average improvement was 4.3 sec (2.2%) in 1.000 m road racing (n = 9); 2.3 sec (1.8%) in 800 m track racing (n = 5); 0.9 sec (1.4%) in 100 m swimming (n = 12); 1.3 sec (0.8%) in 200 m swimming (n = 6); and 2.5 sec (0.9%) in 400 m swimming (n = 3). In a bicycle ergometer test with stepwise, progressive exercise to muscular fatigue, the maximal capacity was increased by 9% (n = 4). Local isometric muscle endurance of the elbow flexors (n = 7) and hand grip strengths (n = 11) were not significantly altered. Possible mechanisms involved in the response to TNS are discussed.

  11. 糖尿病患者自我血糖监测%Self blood glucose monitoring in patients with diabetes

    Institute of Scientific and Technical Information of China (English)

    袁东登

    2015-01-01

    Diabetes mellitus is a polygenic inheritance and environment interaction,a metabolic disease characterized by chronic hyperglycemia.The chronic hyperglycemia can cause serious complications,is a major cause of diabetic death and disability.Full health education,insisting correct long-term treatment measures do not happen or delay complications occur.Through self blood glucose monitoring,effective self management,according to the dynamic changes of the disease,increasing the frequency of monitoring,adjusting treatment measures can improve the blood glucose control effect.Strengthening self blood glucose monitoring can avoid the early occurrence of complications,reduce medical expenses of family and social,improve the life quality of patients with diabetes.%目的:糖尿病是因多基因遗传和环境相互作用引起的,以慢性高血糖为特征的代谢性疾病,其所诱发的严重的并发症,是糖尿病致死、致残的主要原因。进行充分的健康宣教,坚持正确的长期治疗措施,不发生或延缓并发症出现。通过自我血糖监测,进行有效的自我管理,根据病情的动态变化,增加监测的频率,调整治疗措施,提高血糖控制效果。加强自我血糖监测,避免并发症的早期发生,减少家庭和社会的医疗费用,改善糖尿病患者的生活质量。

  12. 24-h ambulatory blood pressure monitoring in healthy young adult Anglo, Hispanic, and African-American subjects.

    Science.gov (United States)

    Chase, H P; Garg, S K; Icaza, G; Carmain, J A; Walravens, C F; Marshall, G

    1997-01-01

    The purpose of this study was to compare office and 24-h ambulatory blood pressure (ABP) values for adolescent and young adult males and females of Anglo, Hispanic, and African-American descent. One hundred and eighteen healthy subjects (62 females, 56 males) participated, with an ethnic distribution of 50 Anglo, 32 Hispanic, and 36 African-American subjects. All subjects came to the clinic for height, weight, sitting blood pressure (BP), and to begin 24-h ABP monitoring using the SpaceLabs model 90207 automatic noninvasive monitor. The monitor recorded readings every 0.5 h from 06:00 to 22:00 and every hour at night from 22:00 to 06:00. Office systolic and diastolic BP values were higher for all males compared to all females. Mean 24-h, nighttime, and daytime systolic ABP values were also significantly higher for males compared to females. The 24-h mean and daytime systolic ABP values were significantly different by ethnic groups. The African-American subjects always had the highest readings. Mean 24-h diastolic ABP was also significantly different by ethnic groups, with the African-American subjects being higher than the Anglos or the Hispanics. Diastolic ABP (24-h mean, daytime, and nighttime) values (for all subjects combined) increased gradually and varied significantly with age. This study provides preliminary normative data about ABP in an understudied population (ie, teenagers and young adults of different ethnic backgrounds). It also shows that higher blood pressures are present among males and among subjects of African-American descent in the teenage and young adult population. PMID:9008244

  13. Laser speckle contrast imaging for monitoring changes in microvascular blood flow

    DEFF Research Database (Denmark)

    Ambrus, Rikard; Strandby, Rune B.; Svendsen, Lars Bo;

    2016-01-01

    the reproducibility of the LSCI technique when assessing the intra-abdominal microcirculation of the pig. METHODS: During trial 1, a fish gill arch was mechanically perfused with heparinized fish blood under controlled stepwise-altered flow rates alongside mechanically induced movement artefacts. The microcirculation......BACKGROUND/AIMS: Microvascular blood flow is essential for healing and predicts surgical outcome. The aim of the current study was to investigate the relation between fluxes measured with the laser speckle contrast imaging (LSCI) technique and changes in absolute blood flow. In addition, we studied...... of the fish gill was simultaneously assessed with the LSCI technique. In trial 2, microcirculation was measured in the stomach, liver, and small intestine of 10 pigs by two observers. RESULTS: A linear correlation was observed between flux and volumetric flow. During conditions of no volumetric flow, the high...

  14. Theoretical considerations to optimize transabdominal monitoring of fetal arterial blood oxygenation using pulse oximetry

    Science.gov (United States)

    Zourabian, Anna; Boas, David A.

    2001-06-01

    Pulse oximetry (oxygen saturation monitoring) has markedly improved medical care in many fields, including anesthesiology, intensive care, and newborn intensive care. In obstetrics, fetal heart rate monitoring remains the standard for intrapartum assessment of fetal well being. Fetal oxygen saturation monitoring is a new technique currently under development. It is potentially superior to electronic fetal heart rate monitoring (cardiotocography) because it allows direct assessment of both fetal oxygen status and fetal tissue perfusion. Here we present the analysis for determining the most optimal wavelength selection for pulse oximetry. The wavelengths we chose as the most optimal are: the first in the range of 670-720nm and the second in the range of 825-925nm. Further we discuss the possible systematic errors during our measurements, and their contribution to the obtained saturation results.

  15. Parameter monitoring system of the Dubna Gas-Filled Recoil Separator

    CERN Document Server

    Tsyganov, Y S; Sukhov, A M

    2015-01-01

    PC-based one-crate monitoring and control system of the Dubna Gas Filled Recoil Separator (DGFRS) is considered. It is developed for the long-term experiments at the U400 FLNR cyclotron and is aimed at the synthesis of super heavy nuclei in heavy ion induced complete fusion reactions. Parameters related to: a) beam and cyclotron; b) separator by itself, c) detection system, d) target and entrance window are measured and stored in the protocol file of the experiment. Special attention is paid to generating the alarm signals and implementing further the appropriate procedures

  16. Casual and ambulatory blood pressure monitoring in children with renal scarring

    Directory of Open Access Journals (Sweden)

    Miloševski Gordana

    2005-01-01

    Full Text Available INTRODUCTION Renal scarring is the most common cause of arterial hypertension in children. High blood pressure (BP and microalbuminuria contribute to the progression of chronic renal disease. OBJECTIVE The aims of the study were: to assess BP in children with renal scarring by continuous ambulatory blood pressure measurement (ABPM in comparison to the casual method (CBP, and to determine the correlation between ambulatory blood pressure (ABP and/or casual blood pressure (CBP values and proteinuria in children with renal scarring. METHOD This forward-looking study comprised thirty-five children (26 girls and 9 boys, aged between 3-13 years, 10.4±3.9, X+SD. Blood pressure was measured using the casual method (CBP with a mercury manometer; BP was measured three times and the average was taken as a referent value. ABPM was performed using the oscillometric method with the Space Labs device, model 90207. RESULTS 45.71% of patients were classified as hypertensive by ABPM, while only 22.6% of CBP measurements were above the 95th percentile (p<0.01. "White coat hypertension" was present in 40% of the patients. Non-dipping BP alteration was detected in 37.14% of the patients. CONCLUSION Nocturnal systolic hypertension (systolic non-dipping alteration is very frequent in children with renal scarring. Nocturnal diastolic blood pressure, detectable only via ABPM, is positively correlated with proteinuria and may be an initial sign of the progression of renal scarring. ABPM is more sensitive than CBP in the evaluation of BP in children with renal scarring.

  17. Moderator's view: Ambulatory blood pressure monitoring and home blood pressure for the prognosis, diagnosis and treatment of hypertension in dialysis patients.

    Science.gov (United States)

    Zoccali, Carmine; Tripepi, Rocco; Torino, Claudia; Tripepi, Giovanni; Mallamaci, Francesca

    2015-09-01

    Major health agencies now recommend the systematic application of ambulatory blood pressure monitoring (ABPM) for the diagnosis of hypertension. Given the exceedingly high prevalence of nocturnal hypertension, masked and white coat hypertension and the overt inadequacy of peridialysis (pre-, intra- and post-dialysis) BP measurements, more extensive application of ABPM for the diagnosis of hypertension in dialysis patients would appear logical. In a recent survey performed in NDT Educational, organizational problems and/or cognitive resistance emerged as important factors hindering more extensive application of ABPM and home BP by nephrologists. External validation of observations made in landmark studies in a single institution about hypertension subcategorization by ABPM is urgently needed. Furthermore, apparent cognitive resistance by nephrologists may be justified by the fact that these techniques have been insufficiently tested in the dialysis population for applicability in everyday clinical practice, tolerability, organizational impact and cost-effectiveness. We should be more resolute in abandoning peridialysis measurements for diagnosing and treating hypertension in haemodialysis patients. Home BP is a formidable educational instrument for patient empowerment and self-care, and evidence exists that this technique is superior to peridialysis values to better hypertension control as defined on the basis of ABPM. We should strive to promote more extensive application of home BP monitoring to diagnose and manage hypertension in haemodialysis patients. ABPM with novel, user friendly and better tolerated techniques is to be awaited in the near future.

  18. Remote sensing for gas plume monitoring using state-of-the-art infrared hyperspectral imaging

    Science.gov (United States)

    Hinnrichs, Michele

    1999-02-01

    Under contract to the US Air Force and Navy, Pacific Advanced Technology has developed a very sensitive hyperspectral imaging infrared camera that can perform remote imaging spectro-radiometry. One of the most exciting applications for this technology is in the remote monitoring of gas plume emissions. Pacific Advanced Technology (PAT) currently has the technology available to detect and identify chemical species in gas plumes using a small light weight infrared camera the size of a camcorder. Using this technology as a remote sensor can give advanced warning of hazardous chemical vapors undetectable by the human eye as well as monitor the species concentrations in a gas plume from smoke stack and fugitive leaks. Some of the gas plumes that have been measured and species detected using an IMSS imaging spectrometer are refinery smoke stacks plumes with emission of CO2, CO, SO2, NOx. Low concentration vapor unseen by the human eye that has been imaged and measured is acetone vapor evaporating at room temperature. The PAT hyperspectral imaging sensor is called 'Image Multi-spectral Sensing or IMSS.' The IMSS instrument uses defractive optic technology and exploits the chromatic aberrations of such lenses. Using diffractive optics for both imaging and dispersion allows for a very low cost light weight robust imaging spectrometer. PAT has developed imaging spectrometers that span the spectral range from the visible, midwave infrared (3 to 5 microns) and longwave infrared (8 to 12 microns) with this technology. This paper will present the imaging spectral data that we have collected on various targets with our hyperspectral imaging instruments as will also describe the IMSS approach to imaging spectroscopy.

  19. A real time, wearable ECG and continous blood pressure monitoring system for first responders.

    Science.gov (United States)

    Ribeiro, David M D; Colunas, Marcio F M; Marques, Fabio A Ferreira; Fernandes, Jose M; Cunha, Joao P Silva

    2011-01-01

    The study of stress and fatigue among First Responders is a major step in mitigating this public health problem. Blood pressure, heart rate variability and fatigue related arrhythmia are three of the main "windows" to study stress and fatigue. In this paper we present a wearable medical device, capable of acquiring an electrocardiogram and estimating blood pressure in real time, through a pulse wave transit time approach. The system is based on an existent certified wearable medical device called "Vital Jacket" and is aimed to become a tool to allow cardiologists in studying stress and fatigue among first response professionals. PMID:22255923

  20. An intelligent system for continuous blood pressure monitoring on remote multi-patients in real time

    OpenAIRE

    Marani, Roberto; Perri, Anna Gina

    2012-01-01

    In this paper we present an electronic system to perform a non-invasive measurement of the blood pressure based on the oscillometric method, which does not suffer from the limitations of the well-known auscultatory one. Moreover the proposed system is able to evaluate both the systolic and diastolic blood pressure values and makes use of a microcontroller and a Sallen-Key active filter. With reference to other similar devices, a great improvement of our measurement system is achieved since it...

  1. Effect of remifentanil combined with propofol anesthesia on stress response, blood gas index in patients with laparoscopic cholecystectomy

    Institute of Scientific and Technical Information of China (English)

    Yan-Dong Jiang; Su-Min Zhao

    2016-01-01

    Objective:To observe the effect of remifentanil combined with propofol anesthesia on stress response, blood gas index in patients with laparoscopic cholecystectomy. Methods:A total of 88 patients with laparoscopic cholecystectomy were randomly divided into the observation group (46 cases) and the control group (42 cases). After anesthesia induction,the observation group was given remifentanil combined with propofol intravenous anesthesia for continuous anesthesia, and control group was given isoflurane and propofol. Stress response (TNF-αα, CRP, BG, COGT), blood gas index (PaO2, PCO2, PH value) before surgery, before anesthesia maintained, and after surgery between two groups were compared. Results: Before anesthesia maintained and after surgery, PaO2 and PCO2 levels of the two groups increased with before surgery (P0.05). pH value of the two groups showed no change before and after surgery. Before anesthesia maintained and after surgery, TNF-αα, CRP, BG, COGT of the two groups increased with before surgery (P0.05) at before anesthesia maintained. While to the end of surgery, these index of the control group was significantly higher than that in the observation group (P<0.05). Conclusions:Remifentanil combined with propofol anesthesia can make blood gas index stable as well as isoflurane combined with propofol anesthesia, but inhibit stress response more significantly.

  2. GULF OF MEXICO SEAFLOOR STABILITY AND GAS HYDRATE MONITORING STATION PROJECT

    Energy Technology Data Exchange (ETDEWEB)

    J. Robert Woolsey; Thomas M. McGee; Robin C. Buchannon

    2004-11-01

    The gas hydrates research Consortium (HRC), established and administered at the University if Mississippi's Center for Marine Research and Environmental Technology (CMRET) has been active on many fronts in FY 03. Extension of the original contract through March 2004, has allowed completion of many projects that were incomplete at the end of the original project period due, primarily, to severe weather and difficulties in rescheduling test cruises. The primary objective of the Consortium, to design and emplace a remote sea floor station for the monitoring of gas hydrates in the Gulf of Mexico by the year 2005 remains intact. However, the possibility of levering HRC research off of the Joint Industries Program (JIP) became a possibility that has demanded reevaluation of some of the fundamental assumptions of the station format. These provisions are discussed in Appendix A. Landmark achievements of FY03 include: (1) Continuation of Consortium development with new researchers and additional areas of research contribution being incorporated into the project. During this period, NOAA's National Undersea Research Program's (NURP) National Institute for Undersea Science and Technology (NIUST) became a Consortium funding partner, joining DOE and Minerals Management Service (MMS); (2) Very successful annual and semiannual meetings in Oxford Mississippi in February and September, 2003; (3) Collection of piston cores from MC798 in support of the effort to evaluate the site for possible monitoring station installation; (4) Completion of the site evaluation effort including reports of all localities in the northern Gulf of Mexico where hydrates have been documented or are strongly suspected to exist on the sea floor or in the shallow subsurface; (5) Collection and preliminary evaluation of vent gases and core samples of hydrate from sites in Green Canyon and Mississippi Canyon, northern Gulf of Mexico; (6) Monitoring of gas activity on the sea floor, acoustically

  3. Implementation of Out-of-Office Blood Pressure Monitoring in the Netherlands: From Clinical Guidelines to Patients' Adoption of Innovation.

    Science.gov (United States)

    Carrera, Pricivel M; Lambooij, Mattijs S

    2015-10-01

    Out-of-office blood pressure monitoring is promoted by various clinical guidelines toward properly diagnosing and effectively managing hypertension and engaging the patient in their care process. In the Netherlands, however, the Dutch cardiovascular risk management (CVRM) guidelines do not explicitly prescribe 24-hour ambulatory blood pressure measurement (ABPM) and home BP measurement (HBPM). The aim of this descriptive study was to develop an understanding of patients' and physicians' acceptance and use of out-of-office BP monitoring in the Netherlands given the CVRM recommendations.Three small focus group discussions (FGDs) with patients and 1 FGD with physicians were conducted to explore the mechanisms behind the acceptance and use of out-of-office BP monitoring and reveal real-world challenges that limit the implementation of out-of-office BP monitoring methods. To facilitate the FGDs, an analytical framework based on the technology acceptance model (TAM), the theory of planned behavior and the model of personal computing utilization was developed to guide the FGDs and analysis of the transcriptions of each FGD.ABPM was the out-of-office BP monitoring method prescribed by physicians and used by patients. HBPM was not offered to patients even with patients' feedback of poor tolerance of ABPM. Even as there was little awareness about HBPM among patients, there were a few patients who owned and used sphygmomanometers. Patients professed and seemed to exhibit self-efficacy, whereas physicians had reservations about (all of their) patients' self-efficacy in properly using ABPM. Since negative experience with ABPM impacted patients' acceptance of ABPM, the interaction of factors that determined acceptance and use was found to be dynamic among patients but not for physicians.In reference to the CVRM guidelines, physicians implemented out-of-office BP monitoring but showed a strong preference for ABPM even where there is poor tolerance of the method. We found that

  4. 家庭血压监测对社区高血压病人血压控制的价值探讨%The Change of Blood Pressure in Patients with Hypertension by Home Blood Pressure Monitoring

    Institute of Scientific and Technical Information of China (English)

    张仁汉; 芦娜; 李洪山; 黄超

    2013-01-01

    Objective: To observe the change of blood pressure, by means of home blood pressure monitoring in patients with hypertension. Method: 212 patients with hypertension were included to proceed home blood pressure monitoring for three months. The data of ambulatory blood pressure monitoring before and after the home blood pressure monitoring were recorded and analyzed. Result:After regular home blood pressure monitoring for three months, the 24h-systolic blood pressure, day systolic blood pressure, and night systolic blood pressure were depressed (P<0.05), and the day mean blood pressure and 24h mean blood pressure were lowered significantly(P<0.05). Conclusion:By means of regular home blood pressure monitoring, the total level of blood pressure are controlled lower in patients with hypertension. It is beneficial for improvement of controlling rate of hypertension and decreasing of complication.%  目的:观察家庭自测血压对高血压病人血压变化的影响。方法:212例门诊高血压病人经过家庭自测血压培训,进行规定的家庭血压监测,持续3个月,观察监测前后动态血压参数的变化。结果:与监测前相比,监测3个月后24h收缩压、白天收缩压、夜间收缩压均有明显降低(P<0.05);在监测3个月后白天平均压和24h平均压也有明显降低(P<0.05)。结论:通过规律的家庭自测血压,可使患者整体血压水平降低,从而有利于提高高血压的控制率和降低高血压的并发症。

  5. Application of geoelectric methods for man-caused gas deposit mapping and monitoring

    Science.gov (United States)

    Yakymchuk, M. A.; Levashov, S. P.; Korchagin, I. N.; Syniuk, B. B.

    2009-04-01

    anomaly total area made S=20.7 hectares on 07.10.08, and S=19.7 hectares on 13.10.08 and S=10.5 hectares on 08.11.08. The anomaly intensity has decreased, some local extremum has appeared. All this testifies that there is an intensive degassing process of cross-section upper part through producing wells and the drilled degassing wells. Exclusively important feature of the FSPEF-VERS technology is an operationability(!) the of practical problems solving. For an emergency situation on gas field an operationability of technology has crucial importance. For one day of works only the field staff management has received considerable volume of operative information, allowing in quite proved manner to estimate as accident scales and it possible reasons, and so those threats, which this accident can represent for nearby located settlements. So, the imposing of a sketch-map of distribution of a "man-caused" gas deposit on a map of wells location has shown that this deposit does not extend over field border and, hence, does not represent essential threat for nearby settlements. Technology operationability in a whole and practical experience of repeated measurements testifies about possibility of the FSPEF-VERS methods using for operative carrying out of monitoring character survey. Such monitoring survey can be spent on a field after degassing wells drilling to check the process of gas pump-down from a "man-caused" deposit. Geoelectric researches on an emergency site of field on 08.11.08 and the received thus results practically show efficiency and working capacity of the FSPEF-VERS technology in a monitoring mode. The performed experimental works have shown, that process of gas pump-down from a "man-caused" deposit can be traced in time by the FSPEF-VERS technology. It is expedient to locate the additional degassing wells for definitive elimination of accident consequences with taking into account the data of monitoring works by FSPEF-VERS methods. The experiment results testify of

  6. [Changes gas and electrolyte structure of blood under influence terahertz radiations on frequencies nitrogen oxide 150,176-150,664 GHz in the conditions of stress].

    Science.gov (United States)

    Tsymbal, A A; Kirichuk, V F

    2011-01-01

    Influence terahertz radiations nitrogen oxide frequencies of 150,176 - 150,664 GHz on gas and electrolyte structure of blood of white rats being in condition of sharp stress. It is shown that at 15 minute mode of influence terahertz waves on frequencies nitrogen oxide observe partial restoration of studied indicators gas and electrolyte structure of blood at stress animals. At 30 minute mode of influence the specified waves observe complete recovery of the broken indicators gas and electrolyte blood structure. PMID:21688667

  7. Beyond White Blood Cell Monitoring : Screening in the Initial Phase of Clozapine Therapy

    NARCIS (Netherlands)

    Cohen, Dan; Bogers, Jan P. A. M.; van Dijk, Daniel; Bakker, Bert; Schulte, Peter F. J.

    2012-01-01

    Objective: Clozapine is the preferred option for treatment-resistant schizophrenia. However, since 1975, clozapine has been known to cause agranulocytosis. In the clozapine screening guidelines, white blood cell count is mandatory. In the past 20 years, after its reintroduction, 3 other serious side

  8. A venous outflow method for continuously monitoring cerebral blood flow in the rat.

    Science.gov (United States)

    Morii, S; Ngai, A C; Ko, K R; Winn, H R

    1986-02-01

    We analyzed the retroglenoid venous outflow (VOF) technique in the rat to document the validity of this method of measuring cerebral blood flow (CBF). Stereotypic changes in CBF were obtained with VOF during hypercarbia and hypotension. O2 content of retroglenoid venous blood did not differ significantly from O2 content of blood obtained from the sagittal sinus, suggesting minimal extracerebral contamination of the retroglenoid venous blood. This lack of extracerebral contamination was further analyzed using a double tracer technique (125I-labeled serum albumin, 22Na) that quantitated minimal extracerebral contamination in the retroglenoid vein. CBF measurements were made simultaneously using microsphere and VOF methods, and excellent correlation was found between the two techniques over a wide range of CBF during normoxia, hypoxia, and normoxic hypocarbia and hypercarbia. However, a decrease in the ratio of VOF to microsphere CBF was observed during severe normoxic hypotension (mean arterial pressure = 41 +/- 4 mmHg). VOF represented 18% of total CBF as measured by microsphere method. This study indicates that the retroglenoid outflow technique in rats is a valid method of measuring CBF. PMID:3080902

  9. Blood Glucose Monitoring as a Teaching Tool for Endocrinology: A New Perspective

    Science.gov (United States)

    Moats, Robert K., II

    2009-01-01

    The education of new allied health professionals and nurses in proper endocrine evaluation and care has become critical in recent years, especially considering the greatly increased prevalence of diabetes in adults and children. The evaluation of blood glucose levels in human volunteers over time is a powerful teaching tool for endocrinology that…

  10. Design of an integrated photo detector circuit for laser doppler blood flow monitoring

    NARCIS (Netherlands)

    Nieland, J.; Kranenburg, van H.; Wallinga, H.; Serov, A.; Steenbergen, W.; Mul, de F.F.M.

    1999-01-01

    A method to measure the blood perfusion of tissue is to apply photons to tissue and measure the frequency shift of the Doppler shifted photons.. To avoid the use of fibers, a chip was designed which contains photodetectors and electronic circuitry to amplify the signal. This IC serves as an interfac

  11. Trans-abdominal monitoring of fetal arterial blood oxygenation using pulse oximetry

    Science.gov (United States)

    Zourabian, Anna; Siegel, Andrew M.; Chance, Britton; Ramanujam, Nirmala; Rode, Martha; Boas, David A.

    2000-10-01

    Pulse oximetry (oxygen saturation monitoring) has markedly improved medical care in many fields, including anesthesiology, intensive care, and newborn intensive care. In obstetrics, fetal heart rate monitoring remains the standard for intrapartum assessment of fetal well being. Fetal oxygen saturation monitoring is a new technique currently under development. It is potentially superior to electronic fetal heart rate monitoring (cardiotocography) because it allows direct assessment of both the fetal oxygen status and fetal tissue perfusion. Here we present the analysis for determining the most optimal wavelength selection for pulse oximetry. The wavelengths we chose as the most optimal are the first in the range of 670 - 720 nm and the second in the range of 825 - 925 nm. Further, we discuss the possible systematic errors during our measurements and their contribution to the obtained saturation results. We present feasibility studies for fetal pulse oximetry, monitored noninvasively through the maternal abdomen. Our preliminary experiments show that the fetal pulse can be discriminated from the maternal pulse and thus, in principle, the fetal arterial oxygen saturation can be obtained. We present the methodology for obtaining these data, and discuss the dependence of our measurements on the fetal position with respect to the optode assembly.

  12. Monitoring and modeling wetland chloride concentrations in relationship to oil and gas development

    Science.gov (United States)

    Post van der Burg, Max; Tangen, Brian A.

    2015-01-01

    Extraction of oil and gas via unconventional methods is becoming an important aspect of energy production worldwide. Studying the effects of this development in countries where these technologies are being widely used may provide other countries, where development may be proposed, with some insight in terms of concerns associated with development. A fairly recent expansion of unconventional oil and gas development in North America provides such an opportunity. Rapid increases in energy development in North America have caught the attention of managers and scientists as a potential stressor for wildlife and their habitats. Of particular concern in the Northern Great Plains of the U.S. is the potential for chloride-rich produced water associated with unconventional oil and gas development to alter the water chemistry of wetlands. We describe a landscape scale modeling approach designed to examine the relationship between potential chloride contamination in wetlands and patterns of oil and gas development. We used a spatial Bayesian hierarchical modeling approach to assess multiple models explaining chloride concentrations in wetlands. These models included effects related to oil and gas wells (e.g. age of wells, number of wells) and surficial geology (e.g. glacial till, outwash). We found that the model containing the number of wells and the surficial geology surrounding a wetland best explained variation in chloride concentrations. Our spatial predictions showed regions of localized high chloride concentrations. Given the spatiotemporal variability of regional wetland water chemistry, we do not regard our results as predictions of contamination, but rather as a way to identify locations that may require more intensive sampling or further investigation. We suggest that an approach like the one outlined here could easily be extended to more of an adaptive monitoring approach to answer questions about chloride contamination risk that are of interest to managers.

  13. Leak before break detection-annulus gas monitoring system evolution and operating experience at KGS

    International Nuclear Information System (INIS)

    Full text: Pressurised heavy water reactors (PHWR) at RAPS 1 and 2 and MAPS have provision for detection of pressure tube leak by indirect method. The reactor vessel (calandria) is housed in calandria vault (C/V) filled with air and C/V moisture element indicates the water leak from calandria tube or pressure tube. Further, detection of leak is a cumbersome process. From NAPS onwards, calandria is housed in C/V filled with water, annulus between calandria tube and pressure tube is filled with CO2 and annulus gas monitoring system (AGMS) is provided by design for detection of any pressure tube leak. The design was improved and AGMS for Kaiga 1 and 2 and RAPS 3 and 4 is having re-circulation mode of operation. The design provides for monitoring dew point of annulus gas (CO2) for indicating the leak and later to identify the pressure tube/calandria tube having leak. The paper deals with operating experience of AGMS at Kaiga generating station (KGS). During the commissioning and initial power operation at KGS, problems were encountered in re-circulation mode. These problems were high radiation field near AGMS piping, high temperature on blower body, blower bearing failure and system leaks. Design modifications were carried out for effective performance of the system for detecting leak before break

  14. Monitoring internal corrosion in natural gas pipelines; Monitoracao da corrosao interna em gasodutos

    Energy Technology Data Exchange (ETDEWEB)

    Nobrega, Ana C.V.; Silva, Djalma R.; Pimenta, Gutemberg S. [Universidade Federal do Rio Grande do Norte, Natal, RN (Brazil); Barbosa, Andrea F.F. [PETROBRAS S.A., Rio de Janeiro, RJ (Brazil). Centro de Pesquisas

    2003-07-01

    For susceptibilities to the corrosion of the pipelines and equipment made in carbon steel and used by the natural gas, it makes be necessary to identify the acting corrosive agents and monitoring them along time, controlling failures for internal corrosion. Also, of that process it origins the black powder (solid particles) that can not commit the structural integrity of the equipment, but it can also bring the company other implications very serious, like quality of the sold product, as well as stops due to blockages and wastes for erosion of the equipment. The monitoring methodology and control of the corrosion in field consisted of the use of corrosion test equipment, chemical characterization of samples of black powder and liquids and analysis of the operational data of processes and plants. Like this, it was identified for the gas pipeline in analysis the most responsible parameters for the corrosive action of the fluid, establishing a controlling methodology and operational actions to maintain the corrosion rates at safe levels and structural warranty of the same. (author)

  15. Gas-Monitor Detector for Intense and Pulsed VUV/EUV Free-Electron Laser Radiation

    Science.gov (United States)

    Sorokin, A. A.; Bobashev, S. V.; Feldhaus, J.; Gerth, Ch.; Gottwald, A.; Hahn, U.; Kroth, U.; Richter, M.; Shmaenok, L. A.; Steeg, B.; Tiedtke, K.; Treusch, R.

    2004-05-01

    In the framework of current developments of new powerful VUV and EUV radiation sources, like VUV free-electron-lasers or EUV plasma sources for 13-nm lithography, we developed a gas-monitor detector in order to measure the photon flux of highly intense and extremely pulsed VUV and EUV radiation in absolute terms. The device is based on atomic photoionization of a rare gas at low particle density. Therefore, it is free of degradation and almost transparent, which allows the detector to be used as a continuously working beam-intensity monitor. The extended dynamic range of the detector allowed its calibration with relative standard uncertainties of 4% in the Radiometry Laboratory of the Physikalisch-Technische Bundesanstalt at the electron-storage ring BESSY II in Berlin using spectrally dispersed synchrotron radiation at low photon intensities and its utilization for absolute photon flux measurements of high power sources. In the present contribution, we describe the design of the detector and its application for the characterization of VUV free-electron-laser radiation at the TESLA test facility in Hamburg. By first pulse resolved measurements, a peak power of more than 100 MW at a wavelength of 87 nm was detected.

  16. Noninvasive transcutaneous bionic baroreflex system prevents severe orthostatic hypotension in patients with spinal cord injury.

    Science.gov (United States)

    Yoshida, Masayoshi; Murayama, Yoshinori; Chishaki, Akiko; Sunagawa, Kenji

    2008-01-01

    Central baroreflex failure in patients with spinal cord injury results in serious orthostatic hypotension. We examined if transcutaneous electrical stimulation regulates arterial pressure in those patients. We identified skin regions capable of increasing arterial pressure and determined respective transfer function. Using the transfer function, we designed the feedback regulator (i.e., bionic baroreflex system) to control arterial pressure. Orthostatic stress decreased arterial pressure profoundly. Activation of bionic regulator restored and maintained arterial pressure at pre-specified levels. We conclude that the transcutaneous bionic system is noninvasive and capable of stabilizing arterial pressure in patients with spinal cord injury.

  17. The effect of siesta in parameters of cardiac structure and in interpretation of ambulatory arterial blood pressure monitoring

    Directory of Open Access Journals (Sweden)

    Marco A.M. Gomes

    2000-04-01

    Full Text Available OBJECTIVE: To evaluate the influence of the siesta in ambulatory blood pressure (BP monitoring and in cardiac structure parameters. METHODS: 1940 ambulatory arterial blood pressure monitoring tests were analyzed (Spacelabs 90207, 15/15 minutes from 7:00 to 22:00 hours and 20/20 minutes from 22:01 to 6.59hours and 21% of the records indicated that the person had taken a siesta (263 woman, 52±14 years. The average duration of the siesta was 118±58 minutes. RESULTS: (average ± standard deviation The average of systolic/diastolic pressures during wakefulness, including the napping period, was less than the average for the period not including the siesta (138±16/85±11 vs 139±16/86±11 mmHg, p5%. CONCLUSION: The siesta influenced the heart structure parameters and from a statistical point of view the average of systolic and diastolic pressures and the respective pressure loads of the wakeful period.

  18. Time-Series Analysis of Continuously Monitored Blood Glucose: The Impacts of Geographic and Daily Lifestyle Factors

    Directory of Open Access Journals (Sweden)

    Sean T. Doherty

    2015-01-01

    Full Text Available Type 2 diabetes is known to be associated with environmental, behavioral, and lifestyle factors. However, the actual impacts of these factors on blood glucose (BG variation throughout the day have remained relatively unexplored. Continuous blood glucose monitors combined with human activity tracking technologies afford new opportunities for exploration in a naturalistic setting. Data from a study of 40 patients with diabetes is utilized in this paper, including continuously monitored BG, food/medicine intake, and patient activity/location tracked using global positioning systems over a 4-day period. Standard linear regression and more disaggregated time-series analysis using autoregressive integrated moving average (ARIMA are used to explore patient BG variation throughout the day and over space. The ARIMA models revealed a wide variety of BG correlating factors related to specific activity types, locations (especially those far from home, and travel modes, although the impacts were highly personal. Traditional variables related to food intake and medications were less often significant. Overall, the time-series analysis revealed considerable patient-by-patient variation in the effects of geographic and daily lifestyle factors. We would suggest that maps of BG spatial variation or an interactive messaging system could provide new tools to engage patients and highlight potential risk factors.

  19. Response of CO2 Concentration in Andisol to Rainfall Events by Using Buried Tubing Gas Monitoring System

    Science.gov (United States)

    Endo, Toshifumi; Tokida, Takeshi; Imoto, Hiromi; Nishimura, Taku; Miyazaki, Tsuyoshi

    For the purpose of continuous soil CO2 gas monitoring, gas permeable resins were evaluated. Among polytetrafluoroethylen (PTFE), polyfluoroethylene propylene (PFEP) and silicone, the silicone rubber tube had highest permeability for oxygen gas. Buried Tubing Gas Monitoring System (BT-GMS) consisting of silicone rubber tube connected to a Non-diffuse infrared (NDIR) -CO2 gas and galvanic cell O2 gas sensors were constructed, and buried into an Andisol upland field at a depth of 20cm. Thermo-couples and EC-5 soil moisture sensors were inserted into 10 and 20cm deep layers. Soil CO2 gas concentration, temperature and moisture were continuously monitored for 5 months. Soil CO2 concentration was sensitive to rainfall events and soil moisture change. Responses were keen during summer until early autumn when soil temperature was higher than 20°C. Then, when soil temperature got lower the response tended to be dull. This suggested quick CO2 gas concentration change following a rainfall event was mostly due to enhancement in soil respiration with soil moisture rise.

  20. Gas monitoring during the CO2 back production field test at the Ketzin pilot site

    Science.gov (United States)

    Szizybalski, Alexandra; Zimmer, Martin; Kujawa, Christian; Erzinger, Jörg

    2015-04-01

    fluids before the injection of CO2 showed that CH4, CO2, H2 and N2 are present in the original formation fluid (0.17 mg/l, 0.08 mg/l, 0.14 mg/l, 17.9 mg/l fluid; Morozova et al., 2010). The observed N2 concentrations in the back-produced gas may, therefore, result from the field tests and from injection management or from the original formation fluid. Results of the isotopic measurements are in preparation. Morozova, D., Wandrey, M., Alawi, M., Zimmer, M., Vieth, A., Zettlitzer, M., Wuerdemann, H. (2010): Monitoring of the microbial community composition in saline aquifers during CO2 storage by fluorescence in situ hybridisation. International Journal of Greenhouse Gas Control, Volume 4, Pages 981-989. doi:10.1016/j.ijggc.2009.11.014.

  1. Characterization of evoked tactile sensation in forearm amputees with transcutaneous electrical nerve stimulation

    Science.gov (United States)

    Chai, Guohong; Sui, Xiaohong; Li, Si; He, Longwen; Lan, Ning

    2015-12-01

    Objective. The goal of this study is to characterize the phenomenon of evoked tactile sensation (ETS) on the stump skin of forearm amputees using transcutaneous electrical nerve stimulation (TENS). Approach. We identified the projected finger map (PFM) of ETS on the stump skin in 11 forearm amputees, and compared perceptual attributes of the ETS in nine forearm amputees and eight able-bodied subjects using TENS. The profile of perceptual thresholds at the most sensitive points (MSPs) in each finger-projected area was obtained by modulating current amplitude, pulse width, and frequency of the biphasic, rectangular current stimulus. The long-term stability of the PFM and the perceptual threshold of the ETS were monitored in five forearm amputees for a period of 11 months. Main results. Five finger-specific projection areas can be independently identified on the stump skin of forearm amputees with a relatively long residual stump length. The shape of the PFM was progressively similar to that of the hand with more distal amputation. Similar sensory modalities of touch, pressure, buzz, vibration, and numb below pain sensation could be evoked both in the PFM of the stump skin of amputees and in the normal skin of able-bodied subjects. Sensory thresholds in the normal skin of able-bodied subjects were generally lower than those in the stump skin of forearm amputees, however, both were linearly modulated by current amplitude and pulse width. The variation of the MSPs in the PFM was confined to a small elliptical area with 95% confidence. The perceptual thresholds of thumb-projected areas were found to vary less than 0.99 × 10-2 mA cm-2. Significance. The stable PFM and sensory thresholds of ETS are desirable for a non-invasive neural interface that can feed back finger-specific tactile information from the prosthetic hand to forearm amputees.

  2. Ultra-sensitive method for determination of ethanol in whole blood by headspace capillary gas chromatography with cryogenic oven trapping.

    Science.gov (United States)

    Watanabe-Suzuki, K; Seno, H; Ishii, A; Kumazawa, T; Suzuki, O

    1999-04-30

    We have established an ultra-sensitive method for determination of ethanol in whole blood by headspace capillary gas chromatography (GC) with cryogenic oven trapping. After heating a blood sample containing ethanol and isobutyl alcohol (internal standard, IS) in a 7.0-ml vial at 55 degrees C for 15 min, 5 ml of the headspace vapor was drawn into a glass syringe and injected into a GC port. All vapor was introduced into an Rtx-BAC2 wide-bore capillary column in the splitless mode at -60 degrees C oven temperature to trap entire analytes, and then the oven temperature was programmed up to 240 degrees C for GC measurements with flame ionization detection. The present method gave sharp peaks of ethanol and IS, and low background noise for whole blood samples. The mean partition into the gaseous phase for ethanol and IS was 3.06+/-0.733 and 8.33+/-2.19%, respectively. The calibration curves showed linearity in the range 0.02-5.0 microg/ml whole blood. The detection limit was estimated to be 0.01 microg/ml. The coefficients of intra-day and inter-day variation for spiked ethanol were 8.72 and 9.47%, respectively. Because of the extremely high sensitivity, we could measure low levels of endogenous ethanol in whole blood of subjects without drinking. The concentration of endogenous ethanol measured for 10 subjects under uncontrolled conditions varied from 0 to 0.377 microg/ml (mean, 0.180 microg/ml). Data on the diurnal changes of endogenous ethanol in whole blood of five subjects under strict food control are also presented; they are in accordance with the idea that endogenous blood ethanol is of enteric bacterial origin.

  3. The pharmacist and the management of arterial hypertension: the role of blood pressure monitoring and telemonitoring.

    Science.gov (United States)

    Omboni, Stefano; Sala, Elisa

    2015-02-01

    Randomized controlled trials have documented that a team of health care professionals which includes a physician, a nurse and a community pharmacist may improve the benefit and adherence of anti-hypertensive therapy. If such a health care model relies on blood pressure telemonitoring, it can promote a stronger relationship between health care professionals and patients, and further improve BP control of hypertension. The major benefit of this collaborative approach is to center the patient's management in a tailored way, providing comprehensive and preventive care based on health information technologies. In this review, the authors summarize recent clinical studies that evaluate the role of the community pharmacist in BP measurements, and in hypertension screening and control. The authors also describe the advantages of using blood pressure telemonitoring in home and ambulatory settings to evaluate potential alternatives to primary care in hypertension management.

  4. Long-range open-path greenhouse gas monitoring using mid-infrared laser dispersion spectroscopy

    Science.gov (United States)

    Daghestani, Nart; Brownsword, Richard; Weidmann, Damien

    2015-04-01

    Accurate and sensitive methods of monitoring greenhouse gas (GHG) emission over large areas has become a pressing need to deliver improved estimates of both human-made and natural GHG budgets. These needs relate to a variety of sectors including environmental monitoring, energy, oil and gas industry, waste management, biogenic emission characterization, and leak detection. To address the needs, long-distance open-path laser spectroscopy methods offer significant advantages in terms of temporal resolution, sensitivity, compactness and cost effectiveness. Path-integrated mixing ratio measurements stemming from long open-path laser spectrometers can provide emission mapping when combined with meteorological data and/or through tomographic approaches. Laser absorption spectroscopy is the predominant method of detecting gasses over long integrated path lengths. The development of dispersion spectrometers measuring tiny refractive index changes, rather than optical power transmission, may offer a set of specific advantages1. These include greater immunity to laser power fluctuations, greater dynamic range due to the linearity of dispersion, and ideally a zero baseline signal easing quantitative retrievals of path integrated mixing ratios. Chirped laser dispersion spectrometers (CLaDS) developed for the monitoring of atmospheric methane and carbon dioxide will be presented. Using quantum cascade laser as the source, a minimalistic and compact system operating at 7.8 μm has been developed and demonstrated for the monitoring of atmospheric methane over a 90 meter open path2. Through full instrument modelling and error propagation analysis, precision of 3 ppm.m.Hz-0.5 has been established (one sigma precision for atmospheric methane normalized over a 1 m path and 1 s measurement duration). The system was fully functional in the rain, sleet, and moderate fog. The physical model and system concept of CLaDS can be adapted to any greenhouse gas species. Currently we are

  5. A method for calculating regional cerebral blood flow from emission computed tomography of inert gas concentrations

    DEFF Research Database (Denmark)

    Celsis, P; Goldman, T; Henriksen, L;

    1981-01-01

    of four 1 min pictures and using a fixed value for the brain:blood partition coefficient, lambda. The method is essentially based on the bolus distribution principle, and it allows the estimation of blood flow in ischemic areas. Application to positron emission tomography is discussed....

  6. Screening for and validated quantification of phenethylamine-type designer drugs and mescaline in human blood plasma by gas chromatography/mass spectrometry.

    Science.gov (United States)

    Habrdova, Vilma; Peters, Frank T; Theobald, Denis S; Maurer, Hans H

    2005-06-01

    In recent years, several newer designer drugs of the so-called 2C series such as 2C-D, 2C-E, 2C-P, 2C-B, 2C-I, 2C-T-2, and 2C-T-7 have entered the illicit drug market as recreational drugs. Some fatal intoxications involving 2C-T-7 have been reported. Only scarce data have been published about analyses of these substances in human blood and/or plasma. This paper describes a method for screening and simultaneous quantification of the above-mentioned compounds and their analog mescaline in human blood plasma. The analytes were analyzed by gas chromatography/mass spectrometry in the selected-ion monitoring mode, after mixed-mode solid-phase extraction (HCX) and derivatization with heptafluorobutyric anhydride. The method was fully validated according to international guidelines. Validation data for 2C-T-2 and 2C-T-7 were unacceptable. For all other analytes, the method was linear from 5 to 500 microg/L and the data for accuracy (bias) and precision (coefficient of variation) were within the acceptance limits of +/-15% and <15%, respectively (within +/-20% and <20% near the limit of quantification of 5 microg/L).

  7. Monitoring cerebral blood flow during intracranial operations: an intravenous injection method

    Energy Technology Data Exchange (ETDEWEB)

    Wyper, D.J.; Pickard, J.D.; Acar, U.

    1979-01-01

    A method of measuring cerebral blood flow during neurosurgery is described. It is an adaptation of the standard 133Xe intravenous injection method, with computation restricted to data collected over the first 3 min. In addition to a theoretical presentation of a method and an analysis of the effect of residual 133Xe activity when sequential runs are performed in close approximation, the technique is validated experimentally by comparison with intracarotid 133Xe injection in baboons.

  8. Characterizing workers participating in a worksite wellness health screening program using blood pressure control, self-monitoring, medication adherence, depression, and exercise.

    Science.gov (United States)

    Breaux-Shropshire, Tonya Lynn; Whitt, Lauren; Griffin, Russell L; Shropshire, Angele Trenese; Calhoun, David A

    2014-07-01

    Blood pressure control remains a serious public health issue because hypertension is the most common risk factor for cardiovascular disease. Effective management of hypertension often requires lifestyle modification and medication adherence. The objective of this study was to identify the prevalence of blood pressure control, medication adherence, self-monitoring of blood pressure, depression, and exercise among workers with access to health resources. Faculty and staff (N = 484) from a university and health care institution in the southeastern United States participated in biometric and questionnaire screening. The researchers used initial screening data from this worksite wellness program to describe baseline blood pressure control (< 140/90 mm Hg), self-monitoring of blood pressure, medication adherence, depression, and exercise. Overall, 63% of the workers' blood pressure was controlled; however, 23% of the sample had been prescribed antihypertensive medication to control their blood pressure. Thirty percent of the sample reported practicing blood pressure self-monitoring, 72.2% reported that they exercised, and 22% reported feeling down and depressed. More than half (64.9%) who used prescribed antihypertensive medication reported adherence to these medications. PMID:25000548

  9. Ambulatory instrument for monitoring indirect beat-to-beat blood pressure in superficial temporal artery using volume-compensation method.

    Science.gov (United States)

    Tanaka, S; Yamakoshi, K

    1996-11-01

    A portable instrument, based on a volume-compensation technique, is designed for ambulatory monitoring of indirect beat-to-beat blood pressure (BP) in the superficial temporal artery. The instrument consists of a small disc-type cuff and a portable unit carried by the subject. Several components are integrated in the cuff for applying counter-pressure to the artery, i.e. a reflectance-type photo-plethysmographic sensor for arterial volume detection, a pressure sensor for cuff pressure Pc measurement and a nozzle flapper-type- electro-pneumatic convertor for controlling Pc. The portable unit includes volume servo control circuitry and a microprocessor-based signal-processing and recording unit. This automatically performs all the necessary measurement procedures and stores into a memory IC element the processed systolic, mean and diastolic blood pressure data, together with pulse intervals on a beat-to-beat basis from the servo-controlled Pc (indirectly measured BP waveform). With this instrument, momentary changes in BP during ambulatory situations such as bicycle ergometer exercise and daily activities including motorway driving are successfully recorded. From the results of simultaneous measurement of the subject's posture changes, the effect of posture change on blood pressure, e.g. baroreceptor-cardiac reflex, is also clearly demonstrated.

  10. Gas bubble trauma monitoring and research of juvenile salmonids. 1995 Annual report

    International Nuclear Information System (INIS)

    This report describes laboratory and field monitoring studies of gas bubble trauma (GBT) in migrating juvenile salmonids in the Snake and Columbia rivers. The first chapter describes laboratory studies of the progression of GBT signs leading to mortality and the use of the signs for GBT assessment. The progression and severity of GBT signs in juvenile salmonids exposed to different levels of total dissolved gas (TDG) and temperatures was assessed and quantified. Next, the prevalence, severity, and individual variation of GBT signs was evaluated to attempt to relate them to mortality. Finally, methods for gill examination in fish exposed to high TDG were developed and evaluated. Primary findings were: (1) no single sign of GBT was clearly correlated with mortality, but many GBT signs progressively worsened; (2) both prevalence and severity of GBT signs in several tissues is necessary; (3) bubbles in the lateral line were the earliest sign of GBT, showed progressive worsening, and had low individual variation but may develop poorly during chronic exposures; (4) fin bubbles had high prevalence, progressively worsened, and may be a persistent sign of GBT; and (5) gill bubbles appear to be the proximate cause of death but may only be relevant at high TDG levels and are difficult to examine. Chapter Two describes monitoring results of juvenile salmonids for signs of GBT. Emigrating fish were collected and examined for bubbles in fins and lateral lines. Preliminary findings were: (1) few fish had signs of GBT, but prevalence and severity appeared to increase as fish migrated downstream; (2) there was no apparent correlation between GBT signs in the fins, lateral line, or gills; (3) prevalence and severity of GBT was suggestive of long-term, non-lethal exposure to relatively low level gas supersaturated water; and (4) it appeared that GBT was not a threat to migrating juvenile salmonids. 24 refs., 26 figs., 3 tabs

  11. Commissioning of the KATRIN Raman system for monitoring of the WGTS gas composition

    Energy Technology Data Exchange (ETDEWEB)

    Fischer, Sebastian; Bornschein, Beate; James, Timothy M.; Off, Andreas; Rupp, Simone; Seitz-Moskaliuk, Hendrik; Sturm, Michael; Wecker, Matthias [Karlsruhe Institute of Technology (Germany); Schloesser, Magnus [Universidad Complutense de Madrid (Spain); Karlsruhe Institute of Technology (Germany); Telle, Helmut H. [Universidad Complutense de Madrid (Spain)

    2015-07-01

    The Karlsruhe Tritium Neutrino (KATRIN) Experiment aims at determining the neutrino mass by the investigation of the endpoint energy-region of the tritium β-spectrum. Tritium in its molecular form ({sup 3}H{sub 2}) is injected into the Windowless Gaseous Tritium Source (WGTS) of KATRIN, and thus molecular effects, e.g. ro-vibrational excitations levels, have to be considered in the neutrino mass analysis in order to reach the design sensitivity of 200 meV/c{sup 2} (90% C.L.). As the source gas also contains impurities of the other hydrogen isotopes protium and deuterium - giving rise to different molecular excitation levels - continuous and precise monitoring of the source gas composition is required. Raman spectroscopy is the method of choice for this task as it is an inline and non-contact analysis method. In this talk, results from the recently performed commissioning phase of the KATRIN Raman system are presented: over the course of more than 50 days, consecutive Raman spectra (recorded with acquisition times of 60 s) of circulating tritium gas were acquired and analysed in real-time. In addition, valuable information on the system performance during long-term operation was gained.

  12. Environmental monitoring and bactericidal efficacy of chlorine dioxide gas in a dental office.

    Science.gov (United States)

    Kuroyama, Iwao; Osato, Shigeo; Nakajima, Shigeki; Kubota, Ryoichi; Ogawa, Takahiro

    2010-09-01

    We monitored the quantity of airborne microorganisms at 11 points (points A to K) in a dental office on a routine day of use, and tested the bactericidal efficacy of chlorine dioxide (ClO₂) gas in the dental operatory after consulting hours. Fallen airborne microorganisms were collected under air-conditioning (AC) in the dental office, and under four conditions in the operatory. Specimens of the microbes were cultivated on nutrient and Sabouraud agar media (NAM and SAM). Many colonies were observed at the entrance hall and on the cabinet in a disinfection room in the NAM and SAM tests, respectively, while no colony was observed at the foot position of the operating table and treatment bed, and above the head position of the operating room in the NAM and SAM tests, respectively. In the bactericidal efficacy test using ClO₂ gas, the dental operatory could be kept clean by the use of 4 mg/L-ClO₂ gas in addition to the use of an AC with a plasma filter and the HEPA filter. PMID:20938095

  13. A graphene-based electrochemical device with thermoresponsive microneedles for diabetes monitoring and therapy

    Science.gov (United States)

    Lee, Hyunjae; Choi, Tae Kyu; Lee, Young Bum; Cho, Hye Rim; Ghaffari, Roozbeh; Wang, Liu; Choi, Hyung Jin; Chung, Taek Dong; Lu, Nanshu; Hyeon, Taeghwan; Choi, Seung Hong; Kim, Dae-Hyeong

    2016-06-01

    Owing to its high carrier mobility, conductivity, flexibility and optical transparency, graphene is a versatile material in micro- and macroelectronics. However, the low density of electrochemically active defects in graphene synthesized by chemical vapour deposition limits its application in biosensing. Here, we show that graphene doped with gold and combined with a gold mesh has improved electrochemical activity over bare graphene, sufficient to form a wearable patch for sweat-based diabetes monitoring and feedback therapy. The stretchable device features a serpentine bilayer of gold mesh and gold-doped graphene that forms an efficient electrochemical interface for the stable transfer of electrical signals. The patch consists of a heater, temperature, humidity, glucose and pH sensors and polymeric microneedles that can be thermally activated to deliver drugs transcutaneously. We show that the patch can be thermally actuated to deliver Metformin and reduce blood glucose levels in diabetic mice.

  14. A fence line noble gas monitoring system for nuclear power plants

    International Nuclear Information System (INIS)

    A noble gas monitoring system has been installed at Ontario Power Generations' Pickering Nuclear Generating Station (PNGS) near Toronto, Canada. This monitoring system allows a direct measure of air kerma from external radiation instead of calculating this based on plant emission data and meteorological models. This has resulted in a reduction in the reported effective dose from external radiation by a factor of at least ten. The system consists of nine self-contained units, each with a 7.6 cm x 7.6 cm (3 inch x 3 inch) NaI(Tl) detector that is calibrated for air kerma. The 512-channel gamma ray spectral information is downloaded daily from each unit to a central computer where the data are stored and processed. A spectral stripping procedure is used to remove natural background variations from the spectral windows used to monitor xenon-133 (133Xe), xenon-135 (135Xe), argon-41 (41Ar), and skyshine radiation from the use of radiography sources. Typical monthly minimum detection limits in air kerma are 0.3 nGy for 133Xe, 0.7 nGy for 135Xe, 3 nGy for 41Ar and 2 nGy for skyshine radiation. Based on 9 months of continuous operation, the annualised air kerma due to 133Xe, 135Xe and 41Ar and skyshine radiation were 7 nGy, 8 nGy, 26 nGy and 107 nGy respectively. (author)

  15. Thirty minute transcutaneous electric acupoint stimulation modulates resting state brain activities: a perfusion and BOLD fMRI study.

    Science.gov (United States)

    Jiang, Yin; Hao, Ying; Zhang, Yue; Liu, Jing; Wang, Xiaoying; Han, Jisheng; Fang, Jing; Zhang, Jue; Cui, Cailian

    2012-05-31

    Increasing neuroimaging studies have focused on the sustained after effects of acupuncture, especially for the changes of brain activities in rest. However, short-period stimuli have mostly been chosen in these works. The present study aimed to investigate how the resting state brain activities in healthy subjects were modulated by relatively long-period (30 min) acupuncture, a widely used modality in clinical practice. Transcutaneous electric acupoint stimulation (TEAS) or intermittent minimal TEAS (MTEAS) were given for 30 min to 40 subjects. Functional MRI (fMRI) data were collected including the pre-stimulation resting state and the post-stimulation resting state, using dual-echo arterial spin labeling (ASL) techniques, representing both cerebral blood flow (CBF) signals and blood oxygen-dependent level (BOLD) signals simultaneously. Following 30 min TEAS, but not MTEAS, the mean global CBF decreased, and a significant decrease of regional CBF was observed in SI, insula, STG, MOG and IFG. Functional connectivity analysis showed more secure and spatially extended connectivity of both the DMN and SMN after 30 min TEAS. Our results implied that modulation of the regional brain activities and network connectivity induced by thirty minute TEAS may associate with the acupuncture-related therapeutic effects. Furthermore, the resting state regional CBF quantified by ASL perfusion fMRI may serve as a potential biomarker in future acupuncture studies. PMID:22541167

  16. Research and Development of Information and Communication Technology-based Home Blood Pressure Monitoring from Morning to Nocturnal Hypertension.

    Science.gov (United States)

    Kario, Kazuomi; Tomitani, Naoko; Matsumoto, Yuri; Hamasaki, Haruna; Okawara, Yukie; Kondo, Maiko; Nozue, Ryoko; Yamagata, Hiromi; Okura, Ayako; Hoshide, Satoshi

    2016-01-01

    Asians have specific characteristics of hypertension (HTN) and its relationship with cardiovascular disease. The morning surge in blood pressure (BP) in Asians is more extended, and the association slope between higher BP and the risk for cardiovascular events is steeper in this population than in whites. Thus, 24-hour BP control including at night and in the morning is especially important for Asian patients with HTN. There are 3 components of "perfect 24-hour BP control": the 24-hour BP level, adequate dipping of nocturnal BP (dipper type), and adequate BP variability such as the morning BP surge. The morning BP-guided approach using home BP monitoring (HBPM) is the first step toward perfect 24-hour BP control. After controlling morning HTN, nocturnal HTN is the second target. We have been developing HBPM that can measure nocturnal BP. First, we developed a semiautomatic HBPM device with the function of automatic fixed-interval BP measurement during sleep. In the J-HOP (Japan Morning Surge Home Blood Pressure) study, the largest nationwide home BP cohort, we successfully measured nocturnal home BP using this device with data memory, 3 times during sleep (2, 3, and 4 am), and found that nocturnal home BP is significantly correlated with organ damage independently of office and morning BP values. The second advance was the development of trigger nocturnal BP (TNP) monitoring with an added trigger function that initiates BP measurements when oxygen desaturation falls below a variable threshold continuously monitored by pulse oximetry. TNP can detect the specific nocturnal BP surges triggered by hypoxic episodes in patients with sleep apnea syndrome. We also added the lowest heart rate-trigger function to TNP to detect the "basal nocturnal BP," which is determined by the circulating volume and structural cardiovascular system without any increase in sympathetic tonus. This double TNP is a novel concept for evaluating the pathogenic pressor mechanism of nocturnal BP

  17. Metal-modified and vertically aligned carbon nanotube sensors array for landfill gas monitoring applications

    Energy Technology Data Exchange (ETDEWEB)

    Penza, M; Rossi, R; Alvisi, M [ENEA, Department of Physical Technologies and New Materials, PO Box 51 Br-4, I-72100 Brindisi (Italy); Serra, E, E-mail: michele.penza@enea.it [ENEA, Department of Physical Technologies and New Materials, Via Anguillarese 301, I-00060 Rome (Italy)

    2010-03-12

    Vertically aligned carbon nanotube (CNT) layers were synthesized on Fe-coated low-cost alumina substrates using radio-frequency plasma enhanced chemical vapour deposition (RF-PECVD) technology. A miniaturized CNT-based gas sensor array was developed for monitoring landfill gas (LFG) at a temperature of 150 deg. C. The sensor array was composed of 4 sensing elements with unmodified CNT, and CNT loaded with 5 nm nominally thick sputtered nanoclusters of platinum (Pt), ruthenium (Ru) and silver (Ag). Chemical analysis of multicomponent gas mixtures constituted of CO{sub 2}, CH{sub 4}, H{sub 2}, NH{sub 3}, CO and NO{sub 2} has been performed by the array sensor responses and pattern recognition based on principal component analysis (PCA). The PCA results demonstrate that the metal-decorated and vertically aligned CNT sensor array is able to discriminate the NO{sub 2} presence in the multicomponent mixture LFG. The NO{sub 2} gas detection in the mixture LFG was proved to be very sensitive, e.g.: the CNT:Ru sensor shows a relative change in the resistance of 1.50% and 0.55% for NO{sub 2} concentrations of 3.3 ppm and 330 ppb dispersed in the LFG, respectively, with a wide NO{sub 2} gas concentration range measured from 0.33 to 3.3 ppm, at the sensor temperature of 150 deg. C. The morphology and structure of the CNT networks have been characterized by scanning electron microscopy (SEM), transmission electron microscopy (TEM) and Raman spectroscopy. A forest-like nanostructure of vertically aligned CNT bundles in the multi-walled form appeared with a height of about 10 {mu}m and a single-tube diameter varying in the range of 5-35 nm. The intensity ratio of the Raman spectroscopy D-peak and G-peak indicates the presence of disorder and defects in the CNT networks. The size of the metal (Pt, Ru, Ag) nanoclusters decorating the CNT top surface varies in the range of 5-50 nm. Functional characterization based on electrical charge transfer sensing mechanisms in the metal

  18. Understanding Challenges and Opportunities of Preventive Blood Pressure Self-Monitoring at Home

    DEFF Research Database (Denmark)

    Grönvall, Erik; Verdezoto, Nervo

    2013-01-01

    methods to understand existing challenges and uncover opportunities of self-monitoring technologies to support preventive healthcare activities among older adults. Emerging challenges from our study were: rule complexity for self-measuring, reliability of measurements, interpretation, understanding...... and health awareness, the sharing activity for prevention, motivational factors, the role of the doctor for prevention, older adult’s use of technology, the home as a distributed information space, and visualization of health data. An awareness of these challenges can help designers to develop better tools...... to support people’s preventive self-monitoring needs compared with existing solutions. Furthermore, supporting the active and informed citizen can improve older adult’s care abilities, awareness and activation towards preventive care....

  19. The correlation and level of agreement between end-tidal and blood gas pCO2 in children with respiratory distress: a retrospective analysis

    Directory of Open Access Journals (Sweden)

    Alexander Jamin L

    2009-03-01

    Full Text Available Abstract Background To investigate the correlation and level of agreement between end-tidal carbon dioxide (EtCO2 and blood gas pCO2 in non-intubated children with moderate to severe respiratory distress. Methods Retrospective study of patients admitted to an intermediate care unit (InCU at a tertiary care center over a 20-month period with moderate to severe respiratory distress secondary to asthma, bronchiolitis, or pneumonia. Patients with venous pCO2 (vpCO2 and EtCO2 measurements within 10 minutes of each other were eligible for inclusion. Patients with cardiac disease, chronic pulmonary disease, poor tissue perfusion, or metabolic abnormalities were excluded. Results Eighty EtCO2-vpCO2 paired values were available from 62 patients. The mean ± SD for EtCO2 and vpCO2 was 35.7 ± 10.1 mmHg and 39.4 ± 10.9 mmHg respectively. EtCO2 and vpCO2 values were highly correlated (r = 0.90, p SD between EtCO2 and vpCO2 was -3.68 ± 4.70 mmHg. The 95% level of agreement ranged from -12.88 to +5.53 mmHg. EtCO2 was found to be more accurate when vpCO2 was 35 mmHg or lower. Conclusion EtCO2 is correlated highly with vpCO2 in non-intubated pediatric patients with moderate to severe respiratory distress across respiratory illnesses. Although the level of agreement between the two methods precludes the overall replacement of blood gas evaluation, EtCO2 monitoring remains a useful, continuous, non-invasive measure in the management of non-intubated children with moderate to severe respiratory distress.

  20. 家庭血压监测中国专家共识%Home blood pressure monitoring: a consensus document of China

    Institute of Scientific and Technical Information of China (English)

    中国医师协会高血压专业委员会; 中国高血压联盟; 中华医学会心血管病学分会

    2012-01-01

    自动的电子血压计正在进入家庭.家庭血压监测已成为血压管理的重要组成部分.近年来,美国、欧洲与日本相继制定了家庭血压监测指南.我国家庭血压监测的研究工作正处起步阶段,尚不能完全根据我国相关研究的结果制定指南.随着高血压患者家庭血压压管理意识的迅速提高,许多家庭拥有了自动的电子血压计,同时也在进行家庭血压测量.为了促进家庭血压监测的健康发展,需要对家庭血压监测进行适当的规范.因此,参考国外有关指南并经广泛、深入的讨论,制定本共识文件.就血压计的选择与校准、血压测量的频率(次数)与时间(天数)以及家庭血压的正常值等重要问题提出了建议.%In China, automated blood pressure monitors recently have been available for home use. Home blood pressure monitoring plays a part in the management of hypertension. In the past few years, the American, European and Japanese societies of hypertension published their guidance document for home blood pressure monitoring. In China, there is still limited data on home blood pressure monitoring, which is insufficient to form a guidance document on the basis of the Chinese data. However, a proportion of Chinese hypertensive patients are becoming aware of the importance of home blood pressure monitoring, and measure blood pressure at home using automated blood pressure monitors. There is therefore a need of guidance- for home blood pressure monitoring. With the reference to the above mentioned guidelines of several other societies and after thorough and comprehensive discussions among a group of Chinese experts on hypertension, this consensus document was then developed to put forward recommendations on the selection and calibration of blood pressure measuring devices, the number of readings and days of blood pressure measurement, and the diagnostic threshold of home blood pressure.

  1. Breath acetone monitoring by portable Si:WO{sub 3} gas sensors

    Energy Technology Data Exchange (ETDEWEB)

    Righettoni, Marco; Tricoli, Antonio; Gass, Samuel [Particle Technology Laboratory, Department of Mechanical and Process Engineering ETH Zurich, CH-8092 Zurich (Switzerland); Schmid, Alex; Amann, Anton [Univ.-Clinic for Anesthesia, Innsbruck Medical University, A-6020 Innsbruck (Austria); Breath Research Institute of the Austrian Academy of Sciences, A-6850 Dornbirn (Austria); Pratsinis, Sotiris E., E-mail: sotiris.pratsinis@ptl.mavt.ethz.ch [Particle Technology Laboratory, Department of Mechanical and Process Engineering ETH Zurich, CH-8092 Zurich (Switzerland)

    2012-08-13

    Highlights: Black-Right-Pointing-Pointer Portable sensors were developed and tested for monitoring acetone in the human breath. Black-Right-Pointing-Pointer Acetone concentrations down to 20 ppb were measured with short response times (<30 s). Black-Right-Pointing-Pointer The present sensors were highly selective to acetone over ethanol and water. Black-Right-Pointing-Pointer Sensors were applied to human breath: good agreement with highly sensitive PTR-MS. Black-Right-Pointing-Pointer Tests with people at rest and during physical activity showed the sensor robustness. - Abstract: Breath analysis has the potential for early stage detection and monitoring of illnesses to drastically reduce the corresponding medical diagnostic costs and improve the quality of life of patients suffering from chronic illnesses. In particular, the detection of acetone in the human breath is promising for non-invasive diagnosis and painless monitoring of diabetes (no finger pricking). Here, a portable acetone sensor consisting of flame-deposited and in situ annealed, Si-doped epsilon-WO{sub 3} nanostructured films was developed. The chamber volume was miniaturized while reaction-limited and transport-limited gas flow rates were identified and sensing temperatures were optimized resulting in a low detection limit of acetone ({approx}20 ppb) with short response (10-15 s) and recovery times (35-70 s). Furthermore, the sensor signal (response) was robust against variations of the exhaled breath flow rate facilitating application of these sensors at realistic relative humidities (80-90%) as in the human breath. The acetone content in the breath of test persons was monitored continuously and compared to that of state-of-the-art proton transfer reaction mass spectrometry (PTR-MS). Such portable devices can accurately track breath acetone concentration to become an alternative to more elaborate breath analysis techniques.

  2. Postoperative transcutaneous oxygen measurement in the prediction of delayed wound healing and prosthetic fitting among amputees during rehabilitation. A pilot study.

    Science.gov (United States)

    Yablon, S A; Novick, E S; Jain, S S; Inhoffer, M; Graves, D E

    1995-01-01

    Postoperative assessment of amputation wound healing remains largely subjective in nature, being based on the physician's clinical judgement. These considerations significantly impact on the rehabilitation course, as premature prosthetic fitting may result in wound breakdown. Alternatively, delayed healing may result in prolonged hospital length of stay. Few attempts have been made to correlate objective parameters of limb perfusion with amputation wound healing or prosthetic fitting outcome during the rehabilitation phase of treatment. A pilot study was conducted, in which the transcutaneous oxygen monitor, a noninvasive device measuring transcutaneous partial pressure of oxygen (tcpO2), was applied to the stumps of 11 consecutive above-or below-knee amputees admitted for rehabilitation after amputation. All patients were tested within 1 wk of admission and 45 days of amputation. The treatment team was blinded as to the test results. A direct correlation was observed between wound healing outcome and tcpO2 results (Fisher's exact test [FET], P = 0.03), and no patient with a tcpO2 of rpbi], = -0.835; P = 0.01), delayed prosthetic fitting (rpbi = 0.742; p = 0.01), and poorer wound healing at admission (rpbi = 0.932; P = 0.001). Postoperative tcpO2 measurement may have use in objectively identifying patients at greater risk of delayed wound healing and prosthetic fitting, although further study is warranted.

  3. The utility of serial complete blood count monitoring in patients receiving radiation therapy for localized prostate cancer

    International Nuclear Information System (INIS)

    Purpose: It is standard practice in our department to monitor weekly complete blood counts (CBCs) in patients receiving definitive radiation therapy for prostate cancer. The clinical utility and cost effectiveness of this practice has not been analyzed. Methods and Materials: The charts of all prostate cancer patients treated with radiation therapy between January 1994 and July 1996 at the Veterans Administration Hospital, Philadelphia, PA were reviewed. CBC values were available for 89 patients. Patients received a median dose of 68 Gy using a four-field box technique and megavoltage photons. Whole-pelvic radiotherapy followed by a conedown to the prostate was administered to 29 patients. Fifty-nine patients received radiation to the prostate alone or prostate and seminal vesicles. Fifty-seven patients received concurrent hormonal therapy which included luteinizing hormone-releasing hormone (LHRH) agonist, antiandrogens, or both. Results: No patient experienced a drop in their hemoglobin, white blood cells (WBCs), or platelets below critical nadirs (defined as WBC 3, hemoglobin 3) at any point during treatment. Thirty-one percent, 25%, and 47% of patients experienced declines of > 50 in platelets, > 2 in WBCs, and > 1 in hemoglobin, respectively. Eight clinical, treatment-related, and pathologic parameters were examined as prognostic factors for a drop of > 2 in hemoglobin, > 2 in WBCs, and > 50 in platelets. Only stage (T3 and above) was significantly associated with a > 2 hemoglobin decline. The use of hormonal therapy was associated with a platelet drop of > 50. None of the 8 prognostic factors were associated with decline of > 2 in WBCs. In the urban area surrounding the Philadelphia Veterans Administration Medical Center, the cost of obtaining a CBC is approximately $30. However, if staff time is considered, the cost of obtaining a weekly CBC during prostate cancer radiotherapy approached $400 per patient. Conclusion: These results suggest that weekly

  4. 42 CFR 414.232 - Special payment rules for transcutaneous electrical nerve stimulators (TENS).

    Science.gov (United States)

    2010-10-01

    ... nerve stimulators (TENS). 414.232 Section 414.232 Public Health CENTERS FOR MEDICARE & MEDICAID SERVICES... Special payment rules for transcutaneous electrical nerve stimulators (TENS). (a) General payment rule. Except as provided in paragraph (b) of this section, payment for TENS is made on a purchase basis...

  5. Effects of transcutaneous electrical nerve stimulation (TENS) on memory in elderly with mild cognitive impairment

    NARCIS (Netherlands)

    Luijpen, MW; Swaab, DF; Sergeant, JA; van Dijk, KRA; Scherder, EJA

    2005-01-01

    In previous studies, transcutaneous electrical nerve stimulation (TENS) was shown to have a positive effect on memory in Alzheimer's disease (AD) patients. Moreover, the reported effects appeared to be more beneficial in early stages of Alzheimer's disease compared to later stage intervention. Based

  6. Effects of transcutaneous electrical nerve stimulation (TENS) on memory in elderly with mild cognitive impairment.

    NARCIS (Netherlands)

    Luijpen, M.W.; Swaab, D.F.; Sergeant, J.A.; Dijk, K.R.A.; Scherder, E.J.

    2005-01-01

    In previous studies, transcutaneous electrical nerve stimulation (TENS) was shown to have a positive effect on memory in Alzheimer's disease (AD) patients. Moreover, the reported effects appeared to be more beneficial in early stages of Alzheimer's disease compared to later stage intervention. Based

  7. A novel approach for an ECG electrode integrated into a transcutaneous sensor.

    Science.gov (United States)

    Hölscher, U

    1987-01-01

    The integration of an ECG-electrode into a common transcutaneous sensor allows simple handling and leads to a reduction of the physiological stress of pre-term infants. Furthermore it may allow future replacement of an invasive method to measure the ECG under labour by a non-invasive one.

  8. Transcutaneous Port for Continuous Duodenal Levodopa/Carbidopa Administration in Parkinson's Disease

    NARCIS (Netherlands)

    Meppelink, Anne Marthe; Nyman, Rickard; van Laar, Teus; Drent, Martje; Prins, Ted; Leenders, Klaus Leonhard

    2011-01-01

    Motor fluctuations in Parkinson's disease (PD) can be reduced by intraduodenal infusion of levodopa-carbidopa (Duodopa (R)) via percutaneous endoscopic gastrojejunostomy (PEG). We applied the transcutaneous soft-tissue anchored titanium port (T-port) in 15 PD patients with motor fluctuations; 7 Duod

  9. Accuracy of transcutaneous carbon dioxide tension measurements during cardiopulmonary exercise testing.

    NARCIS (Netherlands)

    Stege, G.; Elshout, F.J.J. van den; Heijdra, Y.F.; Ven, M.J.T. van de; Dekhuijzen, P.N.R.; Vos, P.J.E.

    2009-01-01

    BACKGROUND: Measurements of transcutaneous carbon dioxide tension (PtcCO(2)) with current devices are proven to provide clinically acceptable agreement with measurements of partial arterial carbon dioxide tension (PaCO(2)) in several settings but not during cardiopulmonary exercise testing (CPET). O

  10. Microneedle arrays for the transcutaneous immunization of diphtheria and influenza in BALB/c mice

    NARCIS (Netherlands)

    Ding, Z.; Verbaan, F. J.; Bivas-Benita, M.; Bungener, L.; Huckriede, A.; van den Berg, D. J.; Kersten, G.; Bouwstra, J. A.

    2009-01-01

    Transcutaneous immunization (TCI) is limited by poor permeation of macromolecules across the skin. Microneedle arrays form transient conduits and enhance the transport of vaccine molecules across the skin barrier without pain sensation. Here we investigated in mouse the immune responses after TO usi

  11. Effects of a single dose of terlipressin on transcutaneous oxygen pressures

    DEFF Research Database (Denmark)

    Krag, Aleksander; Møller, Søren; Henriksen, Jens H;

    2010-01-01

    Abstract Objective. Terlipressin (TP) is a potent vasoconstrictor, which is widely used in the treatment of bleeding esophageal varices and the hepatorenal syndrome. Side effects to TP are often related to skin hypoxaemia. The aim of the study was to investigate the transcutaneous oxygen pressures...

  12. Operational benefits obtained by implementing a remote monitoring and control system at CEGAS (Gas Company of Ceara, Brazil)

    Energy Technology Data Exchange (ETDEWEB)

    Almeida, Rui C.O. [Companhia de Gas do Ceara (CEGAS), Fortaleza, CE (Brazil); Zamponha, Rogerio S. [SOFTBRASIL, Sao Paulo, SP (Brazil)

    2009-07-01

    CEGAS (Gas Company of Ceara) recently implemented an automated remote monitoring and control system at the Natural Gas Stations located within the city of Fortaleza. The main purposes to implement the project were to install a better operational platform, allowing CEGAS to analyze its operational conditions and to measure customer's consumption in real time. The data communication infra-structure chosen was GPRS/GSM, due to it's low deployment cost and coverage availability. The first phase of the project comprised 50 vehicular natural gas stations. The project was successfully installed, and became the 1{sup st} project at this type to run efficiently over a GPRS infra-structure in Brazil for gas monitoring, with reliable control and data communication. This document intends to present the reasons that lead CEGAS to invest in such a system, the technology deployed and the benefits achieved. (author)

  13. Air Pollution Monitoring and Use of Nanotechnology Based Solid State Gas Sensors in Greater Cairo Area, Egypt

    Science.gov (United States)

    Ramadan, A. B. A.

    Air pollution is a serious problem in thickly populated and industrialized areas in Egypt, especially in greater Cairo area. Economic growth and industrialization are proceeding at a rapid pace, accompanied by increasing emissions of air polluting sources. Furthermore, though the variety and quantities of polluting sources have increased dramatically, the development of a suitable method for monitoring the pollution causing sources has not followed at the same pace. Environmental impacts of air pollutants have impact on public health, vegetation, material deterioration etc. To prevent or minimize the damage caused by atmospheric pollution, suitable monitoring systems are urgently needed that can rapidly and reliably detect and quantify polluting sources for monitoring by regulating authorities in order to prevent further deterioration of the current pollution levels. Consequently, it is important that the current real-time air quality monitoring system, controlled by the Egyptian Environmental Affairs Agency (EEAA), should be adapted or extended to aid in alleviating this problem. Nanotechnology has been applied to several industrial and domestic fields, for example, applications for gas monitoring systems, gas leak detectors in factories, fire and toxic gas detectors, ventilation control, breath alcohol detectors, and the like. Here we report an application example of studying air quality monitoring based on nanotechnology `solid state gas sensors'. So as to carry out air pollution monitoring over an extensive area, a combination of ground measurements through inexpensive sensors and wireless GIS will be used for this purpose. This portable device, comprising solid state gas sensors integrated to a Personal Digital Assistant (PDA) linked through Bluetooth communication tools and Global Positioning System (GPS), will allow rapid dissemination of information on pollution levels at multiple sites simultaneously.

  14. Use of audible sound for on-line monitoring of gas metal arc welding process

    Directory of Open Access Journals (Sweden)

    Prezelj, J.

    2008-04-01

    Full Text Available In this paper sound generated during the gas-metal arc welding process was studded. Experimental analyses of the acoustic signals have shown that there are two main noise-generating mechanisms, first having impulse form is arc extinction and arc ignition; the second is the arc itself acting as an ionization sound source. The sound signal is used for assessing and monitoring of the welding process, and for prediction of welding process stability and quality. A new algorithm based on the measured welding current was established for the calculation of emitted sound during the welding process. The comparisons have shown that the calculated values are in good agreement with the measured values.

  15. Investigations into sensing characteristics of circular thin-plate electrostatic sensors for gas path monitoring

    Institute of Scientific and Technical Information of China (English)

    Zhongsheng Chen; Xin Tang; Zheng Hu; Yongmin Yang

    2014-01-01

    Circular thin-plate electrostatic sensors are promising in gas path monitoring due to their advantages of non-intrusiveness and easy installation. The spatial sensitivity and filtering effect are two important performance parameters. In this paper, an analytically mathematical model of induced charge on a circular thin-plate electrode is first derived. Then the spatial sensitivity and fil-tering effect of the circular electrostatic sensor are investigated by numerical calculations. Finally, experimental studies are performed to testify the theoretical results. Both theoretical and experimen-tal results demonstrate that circular thin-plate electrostatic sensors act as a low-pass filter in the spa-tial frequency domain, and both the spatial filtering effect and the temporal frequency response characteristics depend strongly on the spatial position and velocity of the charged particle. These conclusions can provide guidelines for the optimal design of circular thin-plate electrostatic sensors.

  16. Real-time electronic monitoring of a pitted and leaking gas gathering pipeline

    Energy Technology Data Exchange (ETDEWEB)

    Asperger, R.G.; Hewitt, P.G.

    1986-08-01

    Hydrogen patch, flush electrical resistance, and flush linear polarization proves wre used with flush coupons to monitor corrosion rates in a pitted and leaking sour gas gathering line. Four inhibitors were evaluated in stopping the leaks. Inhibitor residuals and the amount and ratio of water and condensate in the lines were measured at five locations along the line. The best inhibitor reduced reduced the pit-leak frequency by over a factor of 10. Inhibitor usage rate was optimized using the hydrogen patch current as a measure of the instantaneous corrosion rate. Improper pigging was identified as a cause of corrosion transients. This problem is discussed in relation to the pigging of pipelines in stratified flow where moving fluids are the carriers for continuously injected corrosion inhibitors.

  17. Recent Developments in Monitoring of Gas and Ash in Volcanic Plumes by Remote Sensing Techniques

    Science.gov (United States)

    Arellano, S. R.

    2007-05-01

    Since the 1970s, a growing list of methods for the remote detection and measurement of the composition and dynamics of volcanic plumes has been available for volcanologists and atmospheric scientists. During the last decade of intensive volcanic activity in Ecuador, the use of spectroscopic techniques like COSPEC, DOAS or FTIR has become an important tool in routine volcano monitoring which has resulted in a better understanding of source and path processes related to volcanogenic gas and ash emissions with increasing spatial-temporal resolution capabilities. The most important developments achieved with these techniques include the incorporation of radiative transfer and diffusion modelling in automatic data processing routines. In addition, work is being done to identify and quantitatively estimate the presence of ash by means of UV spectroscopy. The use of these methods allowed us to follow the degassing process of Tungurahua volcano with unprecedented detail. A brief description of these improvements and their results are presented.

  18. Epileptic Seizure Detection and Prediction Based on Continuous Cerebral Blood Flow Monitoring – a Review

    Directory of Open Access Journals (Sweden)

    Senay Tewolde

    2015-01-01

    Full Text Available Epilepsy is the third most common neurological illness, affecting 1% of the world’s population. Despite advances in medicine, about 25 to 30% of the patients do not respond to or cannot tolerate the severe side effects of medical treatment, and surgery is not an option for the majority of patients with epilepsy. The objective of this article is to review the current state of research on seizure detection based on cerebral blood flow (CBF data acquired by thermal diffusion flowmetry (TDF, and CBF-based seizure prediction. A discussion is provided on the applications, advantages, and disadvantages of TDF in detecting and localizing seizure foci, as well as its role in seizure prediction. Also presented are an overview of the present challenges and possible future research directions (along with methodological guidelines of the CBF-based seizure detection and prediction methods.

  19. Expected trace gas and aerosol retrieval accuracy of the Geostationary Environment Monitoring Spectrometer

    Science.gov (United States)

    Jeong, U.; Kim, J.; Liu, X.; Lee, K. H.; Chance, K.; Song, C. H.

    2015-12-01

    The predicted accuracy of the trace gases and aerosol retrievals from the geostationary environment monitoring spectrometer (GEMS) was investigated. The GEMS is one of the first sensors to monitor NO2, SO2, HCHO, O3, and aerosols onboard geostationary earth orbit (GEO) over Asia. Since the GEMS is not launched yet, the simulated measurements and its precision were used in this study. The random and systematic component of the measurement error was estimated based on the instrument design. The atmospheric profiles were obtained from Model for Ozone And Related chemical Tracers (MOZART) simulations and surface reflectances were obtained from climatology of OMI Lambertian equivalent reflectance. The uncertainties of the GEMS trace gas and aerosol products were estimated based on the OE method using the atmospheric profile and surface reflectance. Most of the estimated uncertainties of NO2, HCHO, stratospheric and total O3 products satisfied the user's requirements with sufficient margin. However, about 26% of the estimated uncertainties of SO2 and about 30% of the estimated uncertainties of tropospheric O3 do not meet the required precision. Particularly the estimated uncertainty of SO2 is high in winter, when the emission is strong in East Asia. Further efforts are necessary in order to improve the retrieval accuracy of SO2 and tropospheric O3 in order to reach the scientific goal of GEMS. Random measurement error of GEMS was important for the NO2, SO2, and HCHO retrieval, while both the random and systematic measurement errors were important for the O3 retrievals. The degree of freedom for signal of tropospheric O3 was 0.8 ± 0.2 and that for stratospheric O3 was 2.9 ± 0.5. The estimated uncertainties of the aerosol retrieval from GEMS measurements were predicted to be lower than the required precision for the SZA range of the trace gas retrievals.

  20. Site selection strategy for environmental monitoring in connection with shale-gas exploration: Vale of Pickering, Yorkshire and Fylde, Lancashire

    OpenAIRE

    Smedley, Pauline L.; Ward, Rob S.; Allen, Grant; Baptie, Brian; Daraktchieva, Zornitza; David G. Jones; Jordan, Colm J.; Purvis, Ruth M.; Cigna, Francesca

    2015-01-01

    This report outlines the strategies for site selection adopted as part of a baseline environmental monitoring investigation in connection with shale-gas exploration and development in the Vale of Pickering, North Yorkshire. The project forms an extension to an ongoing baseline investigation being carried out in the Fylde, Lancashire, and the current project incorporates an air-quality monitoring component that was not within the original remit of the Fylde study. The DECC-funded investigation...