WorldWideScience

Sample records for beam ct kv

  1. Quantitative image quality evaluation for kV cone-beam CT-based IGRT

    International Nuclear Information System (INIS)

    Lim, S Y; Zin, Hafiz M

    2017-01-01

    The objective of this study is to quantitatively evaluate the image quality of a kV cone-beam CT-based IGRT system (Elekta, XVI) using two commercial CT image quality phantoms, Catphan-600 and CIRS-062QA. Both phantoms consist of similar image quality test modules (uniformity, CT linearity and spatial resolution) but each phantom has different diameter and test pattern design. Each test module was imaged separately using an optimised cone-beam CT imaging parameter. The quality metrics of the reconstructed images were analysed using algorithms developed with MatLab. The image uniformity and the spatial resolution measured with Catphan were of 4% and 40% greater respectively, compared to those measured with CIRS phantom. The differences were due to the beam scattering and hardening originated from the CIRS phantom holder. The contrast-to-noise ratio (CNR) values measured with CIRS phantom were at least 2% higher than that of Catphan. The diameter of CIRS phantom is smaller and resulted in lower beam attenuation. The quantitative image quality assessment algorithms developed for both phantoms provided a phantom-specific set of reference values for a cone-beam CT imaging system as recommended by AAPM TG-142. Further investigation will be performed to resolve beam hardening issue arising from the CIRS phantom holder. (paper)

  2. Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation

    Science.gov (United States)

    Yang, Yong; Schreibmann, Eduard; Li, Tianfang; Wang, Chuang; Xing, Lei

    2007-02-01

    On-board CBCT images are used to generate patient geometric models to assist patient setup. The image data can also, potentially, be used for dose reconstruction in combination with the fluence maps from treatment plan. Here we evaluate the achievable accuracy in using a kV CBCT for dose calculation. Relative electron density as a function of HU was obtained for both planning CT (pCT) and CBCT using a Catphan-600 calibration phantom. The CBCT calibration stability was monitored weekly for 8 consecutive weeks. A clinical treatment planning system was employed for pCT- and CBCT-based dose calculations and subsequent comparisons. Phantom and patient studies were carried out. In the former study, both Catphan-600 and pelvic phantoms were employed to evaluate the dosimetric performance of the full-fan and half-fan scanning modes. To evaluate the dosimetric influence of motion artefacts commonly seen in CBCT images, the Catphan-600 phantom was scanned with and without cyclic motion using the pCT and CBCT scanners. The doses computed based on the four sets of CT images (pCT and CBCT with/without motion) were compared quantitatively. The patient studies included a lung case and three prostate cases. The lung case was employed to further assess the adverse effect of intra-scan organ motion. Unlike the phantom study, the pCT of a patient is generally acquired at the time of simulation and the anatomy may be different from that of CBCT acquired at the time of treatment delivery because of organ deformation. To tackle the problem, we introduced a set of modified CBCT images (mCBCT) for each patient, which possesses the geometric information of the CBCT but the electronic density distribution mapped from the pCT with the help of a BSpline deformable image registration software. In the patient study, the dose computed with the mCBCT was used as a surrogate of the 'ground truth'. We found that the CBCT electron density calibration curve differs moderately from that of pCT. No

  3. Evaluation of on-board kV cone beam CT (CBCT)-based dose calculation

    International Nuclear Information System (INIS)

    Yang Yong; Schreibmann, Eduard; Li Tianfang; Wang Chuang; Xing Lei

    2007-01-01

    On-board CBCT images are used to generate patient geometric models to assist patient setup. The image data can also, potentially, be used for dose reconstruction in combination with the fluence maps from treatment plan. Here we evaluate the achievable accuracy in using a kV CBCT for dose calculation. Relative electron density as a function of HU was obtained for both planning CT (pCT) and CBCT using a Catphan-600 calibration phantom. The CBCT calibration stability was monitored weekly for 8 consecutive weeks. A clinical treatment planning system was employed for pCT- and CBCT-based dose calculations and subsequent comparisons. Phantom and patient studies were carried out. In the former study, both Catphan-600 and pelvic phantoms were employed to evaluate the dosimetric performance of the full-fan and half-fan scanning modes. To evaluate the dosimetric influence of motion artefacts commonly seen in CBCT images, the Catphan-600 phantom was scanned with and without cyclic motion using the pCT and CBCT scanners. The doses computed based on the four sets of CT images (pCT and CBCT with/without motion) were compared quantitatively. The patient studies included a lung case and three prostate cases. The lung case was employed to further assess the adverse effect of intra-scan organ motion. Unlike the phantom study, the pCT of a patient is generally acquired at the time of simulation and the anatomy may be different from that of CBCT acquired at the time of treatment delivery because of organ deformation. To tackle the problem, we introduced a set of modified CBCT images (mCBCT) for each patient, which possesses the geometric information of the CBCT but the electronic density distribution mapped from the pCT with the help of a BSpline deformable image registration software. In the patient study, the dose computed with the mCBCT was used as a surrogate of the 'ground truth'. We found that the CBCT electron density calibration curve differs moderately from that of pCT. No

  4. Image guided radiotherapy with the Cone Beam CT kV (ElektaTM): Experience of the Leon Berard Centre

    International Nuclear Information System (INIS)

    Pommier, P.; Gassa, F.; Lafay, F.; Claude, L.

    2009-01-01

    Image guide radiotherapy with the Cone Beam CT kV (C.B.C.T.-kV) developed by Elekta has been implemented at the centre Leon Berard in November 2006. The treatment procedure is presented and detailed for prostate cancer I.G.R.T. and non small cell lung cancer (N.S.C.L.C.) stereotactic radiotherapy (S.R.T.). C.B.C.T.-kV is routinely used for S.R.T., selected paediatric cancers, all prostate carcinomas, primitive brain tumours and head and neck cancers that do not require nodes irradiation. Thirty-five to 40 patients are treated within a daily 11-hours period. The general procedure for 3-dimensional images acquisition and their analysis is described. The C.B.C.T.-kV permitted to identify about 10% of prostate cancer patients for whom a positioning with bone-based 2-dimensional images only would have led to an unacceptable dose distribution for at least one session. S.R.T. is now used routinely for inoperable N.S.C.L.C.. The easiness of implementing C.B.C.T.-kV imaging and its expected medical benefit should lead to a rapid diffusion of this technology that is also submitted to prospective and multi centric medico-economical evaluations. (authors)

  5. Evaluation of image quality for different kV cone-beam CT acquisition and reconstruction methods in the head and neck region

    Energy Technology Data Exchange (ETDEWEB)

    Elstroem, Ulrik V.; Muren, Ludvig P. (Dept. of Oncology, Aarhus Univ. Hospital, Aarhus (Denmark); Dept. of Medical Physics, Aarhus Univ. Hospital, Aarhus (Denmark)), e-mail: ulrielst@rm.dk; Petersen, Joergen B. B. (Dept. of Medical Physics, Aarhus Univ. Hospital, Aarhus (Denmark)); Grau, Cai (Dept. of Oncology, Aarhus Univ. Hospital, Aarhus (Denmark))

    2011-08-15

    Purpose. To evaluate the image quality obtained in a standard QA phantom with both clinical and non-clinical cone-beam computed tomography (CBCT) acquisition modes for the head and neck (HN) region as a step towards CBCT-based treatment planning. The impact of deteriorated Hounsfield unit (HU) accuracy was investigated by comparing results from clinical CBCT image reconstructions to those obtained from a pre-clinical scatter correction algorithm. Methods. Five different CBCT acquisition modes on a clinical system for kV CBCT-guided radiotherapy were investigated. Image reconstruction was performed in both standard clinical software and with an experimental reconstruction algorithm with improved beam hardening and scatter correction. Using the Catphan 504 phantom, quantitative measures of HU uniformity, HU verification and linearity, contrast-to-noise ratio (CNR), and spatial resolution using modulation transfer function (MTF) estimation were assessed. To benchmark the CBCT image properties, comparison to standard HN protocols on conventional CT scanners was performed by similar measures. Results. The HU uniformity within a water-equivalent homogeneous region was considerably improved using experimental vs. standard reconstruction, by factors of two for partial scans and four for full scans. Similarly, the amount of capping/cupping artifact was reduced by more than 1.5%. With mode and reconstruction specific HU calibration using seven inhomogeneity inserts comparable HU linearity was observed. CNR was on average 5% higher for experimental reconstruction (scaled with the square-root of dose between modes for both reconstruction methods). Conclusions. Judged on parameters affecting the common diagnostic image properties, improved beam hardening and scatter correction diminishes the difference between CBCT and CT image quality considerably. In the pursuit of CBCT-based treatment adaptation, dedicated imaging protocols may be required

  6. Monitoring Dosimetric Impact of Weight Loss With Kilovoltage (KV) Cone Beam CT (CBCT) During Parotid-Sparing IMRT and Concurrent Chemotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Ho, Kean Fatt, E-mail: hokeanfatt@hotmail.com [Academic Radiation Oncology, The Christie NHS Foundation Trust, Manchester (United Kingdom); Marchant, Tom; Moore, Chris; Webster, Gareth; Rowbottom, Carl [North Western Medical Physics, The Christie NHS Foundation Trust, Manchester (United Kingdom); Penington, Hazel [Wade Radiotherapy Research Centre, The Christie NHS Foundation Trust, Manchester (United Kingdom); Lee, Lip; Yap, Beng; Sykes, Andrew; Slevin, Nick [Department of Clinical Oncology, The Christie NHS Foundation Trust, Manchester (United Kingdom)

    2012-03-01

    Purpose: Parotid-sparing head-and-neck intensity-modulated radiotherapy (IMRT) can reduce long-term xerostomia. However, patients frequently experience weight loss and tumor shrinkage during treatment. We evaluate the use of kilovoltage (kV) cone beam computed tomography (CBCT) for dose monitoring and examine if the dosimetric impact of such changes on the parotid and critical neural structures warrants replanning during treatment. Methods and materials: Ten patients with locally advanced oropharyngeal cancer were treated with contralateral parotid-sparing IMRT concurrently with platinum-based chemotherapy. Mean doses of 65 Gy and 54 Gy were delivered to clinical target volume (CTV)1 and CTV2, respectively, in 30 daily fractions. CBCT was prospectively acquired weekly. Each CBCT was coregistered with the planned isocenter. The spinal cord, brainstem, parotids, larynx, and oral cavity were outlined on each CBCT. Dose distributions were recalculated on the CBCT after correcting the gray scale to provide accurate Hounsfield calibration, using the original IMRT plan configuration. Results: Planned contralateral parotid mean doses were not significantly different to those delivered during treatment (p > 0.1). Ipsilateral and contralateral parotids showed a mean reduction in volume of 29.7% and 28.4%, respectively. There was no significant difference between planned and delivered maximum dose to the brainstem (p = 0.6) or spinal cord (p = 0.2), mean dose to larynx (p = 0.5) and oral cavity (p = 0.8). End-of-treatment mean weight loss was 7.5 kg (8.8% of baseline weight). Despite a {>=}10% weight loss in 5 patients, there was no significant dosimetric change affecting the contralateral parotid and neural structures. Conclusions: Although patient weight loss and parotid volume shrinkage was observed, overall, there was no significant excess dose to the organs at risk. No replanning was felt necessary for this patient cohort, but a larger patient sample will be investigated

  7. SU-E-I-23: A General KV Constrained Optimization of CNR for CT Abdominal Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Weir, V; Zhang, J [University of Kentucky, Lexington, KY (United States)

    2015-06-15

    Purpose: While Tube current modulation has been well accepted for CT dose reduction, kV adjusting in clinical settings is still at its early stage. This is mainly due to the limited kV options of most current CT scanners. kV adjusting can potentially reduce radiation dose and optimize image quality. This study is to optimize CT abdomen imaging acquisition based on the assumption of a continuous kV, with the goal to provide the best contrast to noise ratio (CNR). Methods: For a given dose (CTDIvol) level, the CNRs at different kV and pitches were measured with an ACR GAMMEX phantom. The phantom was scanned in a Siemens Sensation 64 scanner and a GE VCT 64 scanner. A constrained mathematical optimization was used to find the kV which led to the highest CNR for the anatomy and pitch setting. Parametric equations were obtained from polynomial fitting of plots of kVs vs CNRs. A suitable constraint region for optimization was chosen. Subsequent optimization yielded a peak CNR at a particular kV for different collimations and pitch setting. Results: The constrained mathematical optimization approach yields kV of 114.83 and 113.46, with CNRs of 1.27 and 1.11 at the pitch of 1.2 and 1.4, respectively, for the Siemens Sensation 64 scanner with the collimation of 32 x 0.625mm. An optimized kV of 134.25 and 1.51 CNR is obtained for a GE VCT 64 slice scanner with a collimation of 32 x 0.625mm and a pitch of 0.969. At 0.516 pitch and 32 x 0.625 mm an optimized kV of 133.75 and a CNR of 1.14 was found for the GE VCT 64 slice scanner. Conclusion: CNR in CT image acquisition can be further optimized with a continuous kV option instead of current discrete or fixed kV settings. A continuous kV option is a key for individualized CT protocols.

  8. A novel method for megavoltage scatter correction in cone-beam CT acquired concurrent with rotational irradiation

    NARCIS (Netherlands)

    van Herk, Marcel; Ploeger, Lennert; Sonke, Jan-Jakob

    2011-01-01

    Acquisition of cone-beam CT (CBCT) concurrent with VMAT results in scatter of the megavoltage (MV) beam onto the kilovoltage (kV) detector deteriorating CBCT image quality. The aim of this paper is to develop a method to estimate and correct for MV scatter reaching the kV panel. The correction

  9. Assessment of residual error in liver position using kV cone-beam computed tomography for liver cancer high-precision radiation therapy

    International Nuclear Information System (INIS)

    Hawkins, Maria A.; Brock, Kristy K.; Eccles, Cynthia; Moseley, Douglas; Jaffray, David; Dawson, Laura A.

    2006-01-01

    Purpose: To evaluate the residual error in liver position using breath-hold kilovoltage (kV) cone-beam computed tomography (CT) following on-line orthogonal megavoltage (MV) image-guided breath-hold liver cancer conformal radiotherapy. Methods and Materials: Thirteen patients with liver cancer treated with 6-fraction breath-hold conformal radiotherapy were investigated. Before each fraction, orthogonal MV images were obtained during exhale breath-hold, with repositioning for offsets >3 mm, using the diaphragm for cranio-caudal (CC) alignment and vertebral bodies for medial-lateral (ML) and anterior posterior (AP) alignment. After repositioning, repeat orthogonal MV images, orthogonal kV fluoroscopic movies, and kV cone-beam CTs were obtained in exhale breath-hold. The cone-beam CT livers were registered to the planning CT liver to obtain the residual setup error in liver position. Results: After repositioning, 78 orthogonal MV image pairs, 61 orthogonal kV image pairs, and 72 kV cone-beam CT scans were obtained. Population random setup errors (σ) in liver position were 2.7 mm (CC), 2.3 mm (ML), and 3.0 mm (AP), and systematic errors (Σ) were 1.1 mm, 1.9 mm, and 1.3 mm in the superior, medial, and posterior directions. Liver offsets >5 mm were observed in 33% of cases; offsets >10 mm and liver deformation >5 mm were observed in a minority of patients. Conclusions: Liver position after radiation therapy guided with MV orthogonal imaging was within 5 mm of planned position in the majority of patients. kV cone-beam CT image guidance should improve accuracy with reduced dose compared with orthogonal MV image guidance for liver cancer radiation therapy

  10. Cardiac cone-beam CT

    International Nuclear Information System (INIS)

    Manzke, Robert

    2005-01-01

    This doctoral thesis addresses imaging of the heart with retrospectively gated helical cone-beam computed tomography (CT). A thorough review of the CT reconstruction literature is presented in combination with a historic overview of cardiac CT imaging and a brief introduction to other cardiac imaging modalities. The thesis includes a comprehensive chapter about the theory of CT reconstruction, familiarizing the reader with the problem of cone-beam reconstruction. The anatomic and dynamic properties of the heart are outlined and techniques to derive the gating information are reviewed. With the extended cardiac reconstruction (ECR) framework, a new approach is presented for the heart-rate-adaptive gated helical cardiac cone-beam CT reconstruction. Reconstruction assessment criteria such as the temporal resolution, the homogeneity in terms of the cardiac phase, and the smoothness at cycle-to-cycle transitions are developed. Several reconstruction optimization approaches are described: An approach for the heart-rate-adaptive optimization of the temporal resolution is presented. Streak artifacts at cycle-to-cycle transitions can be minimized by using an improved cardiac weighting scheme. The optimal quiescent cardiac phase for the reconstruction can be determined automatically with the motion map technique. Results for all optimization procedures applied to ECR are presented and discussed based on patient and phantom data. The ECR algorithm is analyzed for larger detector arrays of future cone-beam systems throughout an extensive simulation study based on a four-dimensional cardiac CT phantom. The results of the scientific work are summarized and an outlook proposing future directions is given. The presented thesis is available for public download at www.cardiac-ct.net

  11. Evaluation of variation of voltage (kV) absorbed dose in chest CT scans

    International Nuclear Information System (INIS)

    Mendonca, Bruna G.A.; Mourao, Arnaldo P.

    2013-01-01

    Computed tomography (CT) is one of the most important diagnostic techniques images today. The increasing utilization of CT implies a significant increase of population exposure to ionizing radiation. Optimization of practice aims to reduce doses to patients because the image quality is directly related to the diagnosis. You can decrease the amount of dose to the patient, and maintain the quality of the image. There are several parameters that can be manipulated in a CT scan and these parameters can be used to reduce the energy deposited in the patient. Based on this, we analyzed the variation of dose deposited in the lungs, breasts and thyroid, by varying the supply voltage of the tube. Scans of the thorax were performed following the protocol of routine chest with constant and variable current for the same applied voltage. Moreover, a female phantom was used and thermoluminescent dosimeters (TLD-100), model bat, were used to record the specific organ doses. Scans were performed on a GE CT scanner, model 64 Discovery channels. Higher doses were recorded for the voltage of 120 kV with 200 mAs in the lungs (22.46 mGy) and thyroid (32.22 mGy). For scans with automatic mAs, variable between 100 and 440, this same tension contributed to the higher doses. The best examination in terms of the dose that was used with automatic 80 kV mAs, whose lungs and thyroid received lower dose. For the best breast exam was 100 kV. Since the increase in the 80 kV to 100 kV no impact so much the dose deposited in the lungs, it can be concluded that lowering the applied voltage to 100 kV resulted in a reduction in the dose absorbed by the patient. These results can contribute to optimizing scans of the chest computed tomography

  12. SU-F-J-54: Towards Real-Time Volumetric Imaging Using the Treatment Beam and KV Beam

    Energy Technology Data Exchange (ETDEWEB)

    Chen, M; Rozario, T; Liu, A; Jiang, S; Lu, W [UT Southwestern Medical Center, Dallas, TX (United States)

    2016-06-15

    Purpose: Existing real-time imaging uses dual (orthogonal) kV beam fluoroscopies and may result in significant amount of extra radiation to patients, especially for prolonged treatment cases. In addition, kV projections only provide 2D information, which is insufficient for in vivo dose reconstruction. We propose real-time volumetric imaging using prior knowledge of pre-treatment 4D images and real-time 2D transit data of treatment beam and kV beam. Methods: The pre-treatment multi-snapshot volumetric images are used to simulate 2D projections of both the treatment beam and kV beam, respectively, for each treatment field defined by the control point. During radiation delivery, the transit signals acquired by the electronic portal image device (EPID) are processed for every projection and compared with pre-calculation by cross-correlation for phase matching and thus 3D snapshot identification or real-time volumetric imaging. The data processing involves taking logarithmic ratios of EPID signals with respect to the air scan to reduce modeling uncertainties in head scatter fluence and EPID response. Simulated 2D projections are also used to pre-calculate confidence levels in phase matching. Treatment beam projections that have a low confidence level either in pre-calculation or real-time acquisition will trigger kV beams so that complementary information can be exploited. In case both the treatment beam and kV beam return low confidence in phase matching, a predicted phase based on linear regression will be generated. Results: Simulation studies indicated treatment beams provide sufficient confidence in phase matching for most cases. At times of low confidence from treatment beams, kV imaging provides sufficient confidence in phase matching due to its complementary configuration. Conclusion: The proposed real-time volumetric imaging utilizes the treatment beam and triggers kV beams for complementary information when the treatment beam along does not provide sufficient

  13. SU-F-J-54: Towards Real-Time Volumetric Imaging Using the Treatment Beam and KV Beam

    International Nuclear Information System (INIS)

    Chen, M; Rozario, T; Liu, A; Jiang, S; Lu, W

    2016-01-01

    Purpose: Existing real-time imaging uses dual (orthogonal) kV beam fluoroscopies and may result in significant amount of extra radiation to patients, especially for prolonged treatment cases. In addition, kV projections only provide 2D information, which is insufficient for in vivo dose reconstruction. We propose real-time volumetric imaging using prior knowledge of pre-treatment 4D images and real-time 2D transit data of treatment beam and kV beam. Methods: The pre-treatment multi-snapshot volumetric images are used to simulate 2D projections of both the treatment beam and kV beam, respectively, for each treatment field defined by the control point. During radiation delivery, the transit signals acquired by the electronic portal image device (EPID) are processed for every projection and compared with pre-calculation by cross-correlation for phase matching and thus 3D snapshot identification or real-time volumetric imaging. The data processing involves taking logarithmic ratios of EPID signals with respect to the air scan to reduce modeling uncertainties in head scatter fluence and EPID response. Simulated 2D projections are also used to pre-calculate confidence levels in phase matching. Treatment beam projections that have a low confidence level either in pre-calculation or real-time acquisition will trigger kV beams so that complementary information can be exploited. In case both the treatment beam and kV beam return low confidence in phase matching, a predicted phase based on linear regression will be generated. Results: Simulation studies indicated treatment beams provide sufficient confidence in phase matching for most cases. At times of low confidence from treatment beams, kV imaging provides sufficient confidence in phase matching due to its complementary configuration. Conclusion: The proposed real-time volumetric imaging utilizes the treatment beam and triggers kV beams for complementary information when the treatment beam along does not provide sufficient

  14. Enhancement of image quality with a fast iterative scatter and beam hardening correction method for kV CBCT

    Energy Technology Data Exchange (ETDEWEB)

    Reitz, Irmtraud; Hesse, Bernd-Michael; Nill, Simeon; Tuecking, Thomas; Oelfke, Uwe [DKFZ, Heidelberg (Germany)

    2009-07-01

    The problem of the enormous amount of scattered radiation in kV CBCT (kilo voltage cone beam computer tomography) is addressed. Scatter causes undesirable streak- and cup-artifacts and results in a quantitative inaccuracy of reconstructed CT numbers, so that an accurate dose calculation might be impossible. Image contrast is also significantly reduced. Therefore we checked whether an appropriate implementation of the fast iterative scatter correction algorithm we have developed for MV (mega voltage) CBCT reduces the scatter contribution in a kV CBCT as well. This scatter correction method is based on a superposition of pre-calculated Monte Carlo generated pencil beam scatter kernels. The algorithm requires only a system calibration by measuring homogeneous slab phantoms with known water-equivalent thicknesses. In this study we compare scatter corrected CBCT images of several phantoms to the fan beam CT images acquired with a reduced cone angle (a slice-thickness of 14 mm in the isocenter) at the same system. Additional measurements at a different CBCT system were made (different energy spectrum and phantom-to-detector distance) and a first order approach of a fast beam hardening correction will be introduced. The observed, image quality of the scatter corrected CBCT images is comparable concerning resolution, noise and contrast-to-noise ratio to the images acquired in fan beam geometry. Compared to the CBCT without any corrections the contrast of the contrast-and-resolution phantom with scatter correction and additional beam hardening correction is improved by a factor of about 1.5. The reconstructed attenuation coefficients and the CT numbers of the scatter corrected CBCT images are close to the values of the images acquired in fan beam geometry for the most pronounced tissue types. Only for extreme dense tissue types like cortical bone we see a difference in CT numbers of 5.2%, which can be improved to 4.4% with the additional beam hardening correction. Cupping

  15. The PBX-M 80 kV neutral probe beam

    International Nuclear Information System (INIS)

    Kugel, H.W.; Kaita, R.; Gammel, G.M.; Kozub, T.A.; Roberts, D.W.; Coupland, J.R.; Gray, I.L.S.; Mepham, J.W.

    1989-01-01

    The PBX-M Neutral Probe Beam Project was initiated to design, fabricate, acceptance-test, install, and commission an 80 kV, 2.7 A neutral probe beam for use in PBX-M q-profile measurements. The 100 ms, H 0 output beam is modulated at 1 kHz and has a pulse repetition rate of 3 min. The beam has a divergence of 0.5 0 and a focal length of 400 cm. The injection angle can be varied about 25 0 , from near perpendicular to outboard tangential. The beamline is shielded from magnetic fields and interfaces to the existing mechanical, electrical, and control systems of PBX-M. (orig.)

  16. The PBX-M 80 kV neutral probe beam

    Energy Technology Data Exchange (ETDEWEB)

    Kugel, H.W.; Kaita, R.; Gammel, G.M.; Kozub, T.A.; Roberts, D.W. (Princeton Univ., NJ (USA). Plasma Physics Lab.); Coupland, J.R.; Gray, I.L.S.; Mepham, J.W. (UKAEA Culham Lab., Abingdon)

    1989-04-01

    The PBX-M Neutral Probe Beam Project was initiated to design, fabricate, acceptance-test, install, and commission an 80 kV, 2.7 A neutral probe beam for use in PBX-M q-profile measurements. The 100 ms, H/sup 0/ output beam is modulated at 1 kHz and has a pulse repetition rate of 3 min. The beam has a divergence of 0.5/sup 0/ and a focal length of 400 cm. The injection angle can be varied about 25/sup 0/, from near perpendicular to outboard tangential. The beamline is shielded from magnetic fields and interfaces to the existing mechanical, electrical, and control systems of PBX-M. (orig.).

  17. 150-kV, 80-A solid state power supply for neutral beam injection

    International Nuclear Information System (INIS)

    Owren, H.; Baker, W.; Hopkins, D.; Milnes, K.

    1978-08-01

    A 150-kV, 80-A power supply and neutral beam test facility is now operational at the Lawrence Berkeley Laboratory, Berkeley, California. This supply uses banks of 450-V electrolytic capacitors for over 10 6 joules of energy storage. SCR switches control the power flow to the neutral beam accelerator. Turn on and off times of a few microseconds are possible. An auxiliary capacitor bank also uses SCR switches to provide regulation (''flat-topping'') of the main bank output by switching in additional capacitors as the main bank discharges. Air-operated switches are used to connect the main-bank sections in parallel for charging and series or parallel for discharge, depending on the operating voltage desired. A single digital switch sets the desired operating voltage. Filament and arc power supplies are also solid state. With the exception of the suppressor supply which has one vacuum tube the complete neutral beam system uses all solid state components

  18. Beam profile assessment in spectral CT scanners.

    Science.gov (United States)

    Anjomrouz, Marzieh; Shamshad, Muhammad; Panta, Raj K; Broeke, Lieza Vanden; Schleich, Nanette; Atharifard, Ali; Aamir, Raja; Bheesette, Srinidhi; Walsh, Michael F; Goulter, Brian P; Bell, Stephen T; Bateman, Christopher J; Butler, Anthony P H; Butler, Philip H

    2018-03-01

    In this paper, we present a method that uses a combination of experimental and modeled data to assess properties of x-ray beam measured using a small-animal spectral scanner. The spatial properties of the beam profile are characterized by beam profile shape, the angular offset along the rotational axis, and the photon count difference between experimental and modeled data at the central beam axis. Temporal stability of the beam profile is assessed by measuring intra- and interscan count variations. The beam profile assessment method was evaluated on several spectral CT scanners equipped with Medipix3RX-based detectors. On a well-calibrated spectral CT scanner, we measured an integral count error of 0.5%, intrascan count variation of 0.1%, and an interscan count variation of less than 1%. The angular offset of the beam center ranged from 0.8° to 1.6° for the studied spectral CT scanners. We also demonstrate the capability of this method to identify poor performance of the system through analyzing the deviation of the experimental beam profile from the model. This technique can, therefore, aid in monitoring the system performance to obtain a robust spectral CT; providing the reliable quantitative images. Furthermore, the accurate offset parameters of a spectral scanner provided by this method allow us to incorporate a more realistic form of the photon distribution in the polychromatic-based image reconstruction models. Both improvements of the reliability of the system and accuracy of the volume reconstruction result in a better discrimination and quantification of the imaged materials. © 2018 MARS Bioimaging Ltd. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  19. Empirical beam hardening correction (EBHC) for CT

    Energy Technology Data Exchange (ETDEWEB)

    Kyriakou, Yiannis; Meyer, Esther; Prell, Daniel; Kachelriess, Marc [Institute of Medical Physics, University of Erlangen-Nuernberg, 91052 Erlangen (Germany)

    2010-10-15

    Purpose: Due to x-ray beam polychromaticity and scattered radiation, attenuation measurements tend to be underestimated. Cupping and beam hardening artifacts become apparent in the reconstructed CT images. If only one material such as water, for example, is present, these artifacts can be reduced by precorrecting the rawdata. Higher order beam hardening artifacts, as they result when a mixture of materials such as water and bone, or water and bone and iodine is present, require an iterative beam hardening correction where the image is segmented into different materials and those are forward projected to obtain new rawdata. Typically, the forward projection must correctly model the beam polychromaticity and account for all physical effects, including the energy dependence of the assumed materials in the patient, the detector response, and others. We propose a new algorithm that does not require any knowledge about spectra or attenuation coefficients and that does not need to be calibrated. The proposed method corrects beam hardening in single energy CT data. Methods: The only a priori knowledge entering EBHC is the segmentation of the object into different materials. Materials other than water are segmented from the original image, e.g., by using simple thresholding. Then, a (monochromatic) forward projection of these other materials is performed. The measured rawdata and the forward projected material-specific rawdata are monomially combined (e.g., multiplied or squared) and reconstructed to yield a set of correction volumes. These are then linearly combined and added to the original volume. The combination weights are determined to maximize the flatness of the new and corrected volume. EBHC is evaluated using data acquired with a modern cone-beam dual-source spiral CT scanner (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany), with a modern dual-source micro-CT scanner (TomoScope Synergy Twin, CT Imaging GmbH, Erlangen, Germany), and with a modern

  20. Empirical beam hardening correction (EBHC) for CT

    International Nuclear Information System (INIS)

    Kyriakou, Yiannis; Meyer, Esther; Prell, Daniel; Kachelriess, Marc

    2010-01-01

    Purpose: Due to x-ray beam polychromaticity and scattered radiation, attenuation measurements tend to be underestimated. Cupping and beam hardening artifacts become apparent in the reconstructed CT images. If only one material such as water, for example, is present, these artifacts can be reduced by precorrecting the rawdata. Higher order beam hardening artifacts, as they result when a mixture of materials such as water and bone, or water and bone and iodine is present, require an iterative beam hardening correction where the image is segmented into different materials and those are forward projected to obtain new rawdata. Typically, the forward projection must correctly model the beam polychromaticity and account for all physical effects, including the energy dependence of the assumed materials in the patient, the detector response, and others. We propose a new algorithm that does not require any knowledge about spectra or attenuation coefficients and that does not need to be calibrated. The proposed method corrects beam hardening in single energy CT data. Methods: The only a priori knowledge entering EBHC is the segmentation of the object into different materials. Materials other than water are segmented from the original image, e.g., by using simple thresholding. Then, a (monochromatic) forward projection of these other materials is performed. The measured rawdata and the forward projected material-specific rawdata are monomially combined (e.g., multiplied or squared) and reconstructed to yield a set of correction volumes. These are then linearly combined and added to the original volume. The combination weights are determined to maximize the flatness of the new and corrected volume. EBHC is evaluated using data acquired with a modern cone-beam dual-source spiral CT scanner (Somatom Definition Flash, Siemens Healthcare, Forchheim, Germany), with a modern dual-source micro-CT scanner (TomoScope Synergy Twin, CT Imaging GmbH, Erlangen, Germany), and with a modern

  1. Cone beam CT, wat moet ik ermee?

    NARCIS (Netherlands)

    Hoogeveen, R.

    2013-01-01

    De cone beam-ct-scan (cbct-scan) maakt een opmars in de tandheelkunde vanwege de toegevoegde waarde van de derde dimensie in de diagnostiek. Deze extra informatie wordt verkregen ten koste van een hogere stralenbelasting en een daarmee gepaard gaand hoger risico voor de patiënt. Om de clinicus te

  2. Long-pulse neutral beam power supply system for LBL 20 kV, 10 A sources

    International Nuclear Information System (INIS)

    Honey, V.J.; Baker, W.R.; Fitzgerald, M.L.

    1976-05-01

    A description is given of the power supplies and control system for the LBL 20 kV, 10 A, 10 sec long-pulse neutral beam source test facility, now in operation. Such sources are used in a number of existing and planned fusion power experiments

  3. Study on broad beam heavy ion CT

    International Nuclear Information System (INIS)

    Ohno, Yumiko; Kohno, Toshiyuki; Sasaki, Hitomi; Nanbu, S.; Kanai, Tatsuaki

    2003-01-01

    To achieve the heavy ion radiotherapy more precisely, it is important to know the distribution of the electron density in a human body, which is highly related to the range of charged particles. From a heavy ion CT image, we can directly obtain the 2-D distribution of the electron density in a sample. For this purpose, we have developed a broad beam heavy ion CT system. The electron density was obtained using some kinds of solutions targets. Also the dependence of the spatial resolution on the target size and the kinds of beams was estimated in this work using cylinders targets of 40, 60 and 80 mm in diameter, each of them has a hole of 10 mm in diameter at the center of it. (author)

  4. Beam property measurement of a 300-kV ion source test stand for a 1-MV electrostatic accelerator

    Science.gov (United States)

    Park, Sae-Hoon; Kim, Dae-Il; Kim, Yu-Seok

    2016-09-01

    The KOMAC (Korea Multi-purpose Accelerator Complex) has been developing a 300-kV ion source test stand for a 1-MV electrostatic accelerator for industrial purposes. A RF ion source was operated at 200 MHz with its matching circuit. The beam profile and emittance were measured behind an accelerating column to confirm the beam property from the RF ion source. The beam profile was measured at the end of the accelerating tube and at the beam dump by using a beam profile monitor (BPM) and wire scanner. An Allison-type emittance scanner was installed behind the beam profile monitor (BPM) to measure the beam density in phase space. The measurement results for the beam profile and emittance are presented in this paper.

  5. Scatter corrections for cone beam optical CT

    Energy Technology Data Exchange (ETDEWEB)

    Olding, Tim; Holmes, Oliver [Department of Physics, Queen' s University (United Kingdom); Schreiner, L John [Medical Physics Department, Cancer Centre of Southeastern Ontario (Canada)], E-mail: Tim.Olding@krcc.on.ca

    2009-05-01

    Cone beam optical computed tomography (OptCT) employing the VISTA scanner (Modus Medical, London, ON) has been shown to have significant promise for fast, three dimensional imaging of polymer gel dosimeters. One distinct challenge with this approach arises from the combination of the cone beam geometry, a diffuse light source, and the scattering polymer gel media, which all contribute scatter signal that perturbs the accuracy of the scanner. Beam stop array (BSA), beam pass array (BPA) and anti-scatter polarizer correction methodologies have been employed to remove scatter signal from OptCT data. These approaches are investigated through the use of well-characterized phantom scattering solutions and irradiated polymer gel dosimeters. BSA corrected scatter solutions show good agreement in attenuation coefficient with the optically absorbing dye solutions, with considerable reduction of scatter-induced cupping artifact at high scattering concentrations. The application of BSA scatter corrections to a polymer gel dosimeter lead to an overall improvement in the number of pixel satisfying the (3%, 3mm) gamma value criteria from 7.8% to 0.15%.

  6. Dose reduction in dynamic CT stress myocardial perfusion imaging: comparison of 80-kV/370-mAs and 100-kV/300-mAs protocols.

    Science.gov (United States)

    Fujita, Makiko; Kitagawa, Kakuya; Ito, Tatsuro; Shiraishi, Yasuyuki; Kurobe, Yusuke; Nagata, Motonori; Ishida, Masaki; Sakuma, Hajime

    2014-03-01

    To determine the effect of reduced 80-kV tube voltage with increased 370-mAs tube current on radiation dose, image quality and estimated myocardial blood flow (MBF) of dynamic CT stress myocardial perfusion imaging (CTP) in patients with a normal body mass index (BMI) compared with a 100-kV and 300-mAs protocol. Thirty patients with a normal BMI (stress dual-source dynamic CTP. Patients were randomised to 80-kV/370-mAs (n = 15) or 100-kV/300-mAs (n = 15) imaging. Maximal enhancement and noise of the left ventricular (LV) cavity, contrast-to-noise ratio (CNR) and MBF of the two groups were compared. Imaging with 80-kV/370-mAs instead of 100-kV/300-mAs was associated with 40% lower radiation dose (mean dose-length product, 359 ± 66 vs 628 ± 112 mGy[Symbol: see text]cm; P noise (22.7 ± 3.5 vs 17.4 ± 2.6; P stress perfusion imaging (CTP) is increasingly used to assess myocardial function. • Dynamic CTP is feasible at 80-kV in patients with normal BMI. • An 80-kV/370-mAs protocol allows 40% dose reduction compared with 100-kV/300-mAs. • Contrast-to-noise ratio and myocardial blood flow of the two protocols were comparable.

  7. Design and test of a 40-kV, 80-A, 10-msec, neutral-beam power supply series

    International Nuclear Information System (INIS)

    North, G.G.

    1977-01-01

    To meet neutral-beam source requirements, a combination series switch/regulator system has been developed that can provide up to 40-kV at 80A output for 10-ms from the continuously decaying voltage of a charged capacitor bank. The system uses 100% feedback control of a series hard tube regulator. This feedback regulator is able to maintain a 40-kV output level for 100% load variations while the source voltage for the capacitor bank is drained from an initial 55-kV down to as low as 43-kV during a 10-ms pulse. In addition to controlling the output voltage, the series regulator tube also serves the dual role of a disconnect or interrupt switch at the end of each pulse and during the frequent occurrence of a neutral-beam source fault. In the interrupt mode, complete disconnect is achieved in less than 2-μs after first observance of a fault condition; recovery times to normal operation of less than 10-μs after fault clearance can be attained if desired

  8. Comparison of CT numbers between cone-beam CT and multi-detector CT

    International Nuclear Information System (INIS)

    Kim, Dong Soo; Han, Won Jeong; Kim, Eun Kyung

    2010-01-01

    To compare the CT numbers on 3 cone-beam CT (CBCT) images with those on multi-detector CT (MDCT) image using CT phantom and to develop linear regressive equations using CT numbers to material density for all the CT scanner each. Mini CT phantom comprised of five 1 inch thick cylindrical models with 1.125 inches diameter of materials with different densities (polyethylene, polystyrene, plastic water, nylon and acrylic) was used. It was scanned in 3 CBCTs (i-CAT, Alphard VEGA, Implagraphy SC) and 1 MDCT (Somatom Emotion). The images were saved as DICOM format and CT numbers were measured using OnDemand 3D. CT numbers obtained from CBCTs and MDCT images were compared and linear regression analysis was performed for the density, ρ(g/cm 3 ), as the dependent variable in terms of the CT numbers obtained from CBCTs and MDCT images. CT numbers on i-CAT and Implagraphy CBCT images were smaller than those on Somatom Emotion MDCT image (p<0.05). Linear relationship on a range of materials used for this study were ρ=0.001 H+1.07 with R2 value of 0.999 for Somatom Emotion, ρ=0.002 H+1.09 with R2 value of 0.991 for Alphard VEGA, ρ=0.001 H+1.43 with R2 value of 0.980 for i-CAT and ρ=0.001 H+1.30 with R2 value of 0.975 for Implagraphy. CT numbers on i-CAT and Implagraphy CBCT images were not same as those on Somatom Emotion MDCT image. The linear regressive equations to determine the density from the CT numbers with very high correlation coefficient were obtained on three CBCT and MDCT scan.

  9. Improvement of image quality and radiation dose of CT perfusion of the brain by means of low-tube voltage (70 KV)

    International Nuclear Information System (INIS)

    Li, Zhen-lin; Zhang, Kai; Li, Wang-jiang; Chen, Xian; Wu, Bin; Song, Bin; Li, Hang

    2014-01-01

    To investigate the feasibility of 70 kV cerebral CT perfusion by comparing image quality and radiation exposure to 80 kV. Thirty patients with suspected cerebral ischemia who underwent dual-source CT perfusion were divided into group A (80 kV, 150 mAs) and group B (70 kV, 150 mAs). Quantitative comparisons were used for maximum enhancement, signal-to-noise index (SNI), and values of cerebral blood flow (CBF), cerebral blood flow (CBV), mean transit time (MTT) on CBF, CBV, and MTT images, and radiation dose from these two groups. Qualitative perfusion images were assessed by two readers. Maximum enhancement for group B was higher than group A (P < 0.05). There were no significant differences between the two groups for SNI on CBF and CBV maps (P = 0.06 - 0.576), but significant differences for MTT when SNI was measured on frontal white matter and temporo-occipital white matter (P < 0.05). There were no differences among values of CBF, CBV, and MTT for both groups (P = 0.251-0.917). Mean image quality score in group B was higher than group A for CBF (P < 0.05), but no differences for CBV (P = 0.542) and MTT (P = 0.962). Radiation dose for group B decreased compared with group A. 70 kV cerebral CT perfusion reduces radiation dose without compromising image quality. (orig.)

  10. [Evaluation of low contrast resolution in cone beam CT using FPD].

    Science.gov (United States)

    Sakamoto, Kiyoshi; Miura, Koui; Ueda, Ken; Baba, Rika; Kamakura, Toshiko; Sakamoto, Rieko; Okabe, Masakazu; Nakao, Norio

    2006-04-20

    We examined the low contrast resolution of cone beam CT (CBCT) equipped with an indirect-type flat panel detector and compared it with a commercial CT unit (Robusto) . In CBCT, the X-ray tube voltage of 110 kV was used, and in the Robusto, the usual 120 kV was used for examinations. The computed tomography dose index (CTDI) of the two systems was measured, and images scanned at about the same exposure to radiation were compared. The modulation transfer factors of the two systems were measured, and the convolution kernel that was the nearest to the characteristic of CBCT was chosen among kernels of the Robusto. A water phantom with a diameter of 200 mm was scanned, Wiener spectra were calculated, and signal-to-noise ratios were compared. The low contrast resolution phantom was scanned, and detectability and contrast-to-noise ratio (CNR) were measured. In addition, we placed diluted contrast medium into a phantom, scanned the phantom, and measured the detectability and CNR. When the X-ray irradiation condition of CBCT was 75 mAs at 110 kV, the equal dose of radioactivity in the Robusto was 50 mAs at 120 kV. In the low contrast resolution phantom, detectability was 8.7%mm in CBCT, and 9.4%mm in the Robusto. In the low contrast resolution evaluation phantom, CNR was 1.39 in CBCT, and 2.69 in the Robusto. With diluted contrast medium, CNR was 1.28 in CBCT, and 0.60 in the Robusto. CBCT was inferior to the Robusto in a low contrast resolution phantom, but CBCT was superior to the Robusto using diluted contrast medium. We found that CBCT was useful in examinations using contrast media.

  11. On-line image guided radiation therapy using cone-beam CT (CBCT)

    International Nuclear Information System (INIS)

    Bak, Jin O; Park, Suk Won; Jeong, Kyung Keun; Keum, Ki Chang

    2006-01-01

    Using cone beam CT, we can compare the position of the patients at the simulation and the treatment. In on-line image guided radiation therapy, one can utilize this compared data and correct the patient position before treatments. Using cone beam CT, we investigated the errors induced by setting up the patients when use only the markings on the patients' skin. We obtained the data of three patients that received radiation therapy at the Department of Radiation Oncology in Chung-Ang University during August 2006 and October 2006. Just as normal radiation therapy, patients were aligned on the treatment couch after the simulation and treatment planning. Patients were aligned with lasers according to the marking on the skin that were marked at the simulation time and then cone beam CTs were obtained. Cone beam CTs were fused and compared with simulation CTs and the displacement vectors were calculated. Treatment couches were adjusted according to the displacement vector before treatments. After the treatment, positions were verified with kV X-ray (OBI system). In the case of head and neck patients, the average sizes of the setup error vectors, given by the cone beam CT, were 0.19 cm for the patient A and 0.18 cm for the patient B. The standard deviations were 0.15 cm and 0.21 cm, each. On the other hand, in the case of the pelvis patient, the average and the standard deviation were 0.37 cm and 01 cm. Through the on-line IGRT using cone beam CT, were could correct the setup errors that could occur in the conventional radiotherapy. The importance of the on-line IGRT should be emphasized in the case of 3D conformal therapy and intensity-modulated radiotherapy, which have complex target shapes and steep dose gradients

  12. Prospective ECG-triggered axial CT at 140-kV tube voltage improves coronary in-stent restenosis visibility at a lower radiation dose compared with conventional retrospective ECG-gated helical CT

    Energy Technology Data Exchange (ETDEWEB)

    Horiguchi, Jun; Fujioka, Chikako; Kiguchi, Masao; Kohno, Shingo [Hiroshima University Hospital, Department of Clinical Radiology, Hiroshima (Japan); Yamamoto, Hideya; Kitagawa, Toshiro [Hiroshima University, Department of Molecular and Internal Medicine, Division of Clinical Medical Science, Programs for Applied Biomedicine, Graduate School of Biomedical Sciences, Hiroshima (Japan); Ito, Katsuhide [Hiroshima University, Department of Radiology, Division of Medical Intelligence and Informatics, Programs for Applied Biomedicine, Graduate School of Biomedical Sciences, Hiroshima (Japan)

    2009-10-15

    The purpose of this study was to compare coronary 64-slice CT angiography (CTA) protocols, specifically prospective electrocardiograph (ECG)-triggered and retrospective ECG-gated CT acquisition performed using a tube voltage of 140 kV and 120 kV, regarding intracoronary stent imaging. Coronary artery stents (n=12) with artificial in-stent restenosis (50% luminal reduction, 40 HU) on a cardiac phantom were examined by CT at heart rates of 50-75 beats per minute (bpm). The subjective visibility of in-stent restenosis was evaluated with a three-point scale (1 clearly visible, 2 visible, and 3 not visible), and artificial lumen narrowing [(inner stent diameter - measured lumen diameter)/inner stent diameter], lumen attenuation increase ratio [(in-stent attenuation - coronary lumen attenuation)/coronary lumen attenuation], and signal-to-noise ratio of in-stent lumen were determined. The effective dose was estimated. The artificial lumen narrowing (mean 43%), the increase of lumen attenuation (mean 46%), and signal-to-noise ratio (mean 7.8) were not different between CT acquisitions (p=0.12-0.91). However, the visibility scores of in-stent restenosis were different (p<0.05) between ECG-gated CTA techniques: (a) 140-kV prospective (effective dose 4.6 mSv), 1.6; (b) 120-kV prospective (3.3 mSv), 1.8; (c) 140-kV retrospective (16.4-18.8 mSv), 1.9; and (d) 120-kV retrospective (11.0-13.4 mSv), 1.9. Thus, 140-kV prospective ECG-triggered CTA improves coronary in-stent restenosis visibility at a lower radiation dose compared with retrospective ECG-gated CTA. (orig.)

  13. Pediatric thoracic CT angiography at 70 kV: a phantom study to investigate the effects on image quality and radiation dose

    International Nuclear Information System (INIS)

    MacDougall, Robert D.; Kleinman, Patricia L.; Lee, Edward Y.; Yu, Lifeng

    2016-01-01

    Studies have demonstrated that 70-kilovolt (kV) imaging enhances the contrast of iodine, potentially affording a reduction in radiation dose while maintaining the contrast-to-noise ratio (CNR). There is a maximum amount of image noise beyond which increased contrast does not improve structure visualization. Thus, noise should be constrained during protocol optimization. This phantom study investigated the effect of 70-kV imaging for pediatric thoracic CT angiography on image quality and radiation dose in a pediatric population when a noise constraint was considered. We measured contrast and noise using anthropomorphic thoracic phantoms ranging in size from newborn age equivalent to 10-year-old age equivalent. We inserted contrast rods into the phantoms to simulate injected contrast material used in a CT angiography study. The image-quality metric ''iodine CNR with a noise constraint'' was used to determine the relative dose factor for each phantom size, kV setting (70-140 kV) and noise constraint (1.00-1.20). A noise constraint of 1.20 indicates that noise should not increase by more than 20% of the noise level in images performed at the reference kV, selected to be 80 kV in this study. The relative dose factor can be applied to the original dose obtained at 80 kV in order to maintain iodine CNR with the noise constraint. A relative dose factor <1.0 indicates potential for dose reduction while a relative dose factor >1.0 indicates a dose penalty. Iodine contrast was highest for 70 kV and decreased with higher kV settings for all phantom sizes. The relative dose factor at 70 kV was <1.0 for all noise constraint >1.0, indicating potential for dose reduction, for the newborn, 1-year-old and 5-year-old age-equivalent phantom sizes. For the 10-year-old age-equivalent phantom, relative dose factor at 70 kV=1.22, 1.11, 1.01, 0.92 and 0.83 for noise constraint=1.00, 1.05, 1.10, 1.15, 1.20, respectively, indicating a dose penalty for noise constraint

  14. Computer code determination of tolerable accel current and voltage limits during startup of an 80 kV MFTF sustaining neutral beam source

    International Nuclear Information System (INIS)

    Mayhall, D.J.; Eckard, R.D.

    1979-01-01

    We have used a Lawrence Livermore Laboratory (LLL) version of the WOLF ion source extractor design computer code to determine tolerable accel current and voltage limits during startup of a prototype 80 kV Mirror Fusion Test Facility (MFTF) sustaining neutral beam source. Arc current limits are also estimated. The source extractor has gaps of 0.236, 0.721, and 0.155 cm. The effective ion mass is 2.77 AMU. The measured optimum accel current density is 0.266 A/cm 2 . The gradient grid electrode runs at 5/6 V/sub a/ (accel voltage). The suppressor electrode voltage is zero for V/sub a/ < 3 kV and -3 kV for V/sub a/ greater than or equal to 3 kV. The accel current density for optimum beam divergence is obtained for 1 less than or equal to V/sub a/ less than or equal to 80 kV, as are the beam divergence and emittance

  15. Cone Beam CT vs. Fan Beam CT: A Comparison of Image Quality and Dose Delivered Between Two Differing CT Imaging Modalities.

    Science.gov (United States)

    Lechuga, Lawrence; Weidlich, Georg A

    2016-09-12

    A comparison of image quality and dose delivered between two differing computed tomography (CT) imaging modalities-fan beam and cone beam-was performed. A literature review of quantitative analyses for various image quality aspects such as uniformity, signal-to-noise ratio, artifact presence, spatial resolution, modulation transfer function (MTF), and low contrast resolution was generated. With these aspects quantified, cone beam computed tomography (CBCT) shows a superior spatial resolution to that of fan beam, while fan beam shows a greater ability to produce clear and anatomically correct images with better soft tissue differentiation. The results indicate that fan beam CT produces superior images to that of on-board imaging (OBI) cone beam CT systems, while providing a considerably less dose to the patient.

  16. Topogram-based automated selection of the tube potential and current in thoraco-abdominal trauma CT - a comparison to fixed kV with mAs modulation alone

    Energy Technology Data Exchange (ETDEWEB)

    Frellesen, Claudia; Stock, Wenzel; Kerl, J.M.; Lehnert, Thomas; Wichmann, Julian L.; Beeres, Martin; Schulz, Boris; Bodelle, Boris; Vogl, Thomas J. [Clinic of the Goethe University, Department of Diagnostic and Interventional Radiology, Frankfurt (Germany); Nau, Christoph; Geiger, Emanuel; Wutzler, Sebastian [Clinic of the Goethe University, Department of Trauma, Hand and Reconstructive Surgery, Frankfurt (Germany); Ackermann, Hanns [Clinic of the Goethe University, Department of Biostatistics and Mathematical Modelling, Frankfurt (Germany); Bauer, Ralf W. [Clinic of the Goethe University, Department of Diagnostic and Interventional Radiology, Frankfurt (Germany); Klinikum der Goethe-Universitaet, Institut fuer Diagnostische und Interventionelle Radiologie, Frankfurt am Main (Germany)

    2014-07-15

    To investigate the impact of automated attenuation-based tube potential selection on image quality and exposure parameters in polytrauma patients undergoing contrast-enhanced thoraco-abdominal CT. One hundred patients were examined on a 16-slice device at 120 kV with 190 ref.mAs and automated mA modulation only. Another 100 patients underwent 128-slice CT with automated mA modulation and topogram-based automated tube potential selection (autokV) at 100, 120 or 140 kV. Volume CT dose index (CTDI{sub vol}), dose-length product (DLP), body diameters, noise, signal-to-noise ratio (SNR) and subjective image quality were compared. In the autokV group, 100 kV was automatically selected in 82 patients, 120 kV in 12 patients and 140 kV in 6 patients. Patient diameters increased with higher kV settings. The median CTDI{sub vol} (8.3 vs. 12.4 mGy; -33 %) and DLP (594 vs. 909 mGy cm; -35 %) in the entire autokV group were significantly lower than in the group with fixed 120 kV (p < 0.05 for both). Image quality remained at a constantly high level at any selected kV level. Topogram-based automated selection of the tube potential allows for significant dose savings in thoraco-abdominal trauma CT while image quality remains at a constantly high level. (orig.)

  17. The Utrecht 850 kV cascade generator I. Beam deflection and energy control

    NARCIS (Netherlands)

    Braams, C.M.; Smith, P.B.

    The beam deflection magnet and energy control system of the Utrecht cascade generator are described. The uniform-field magnet has entrance and exit slits located outside the magnetic held. Since the cascade generator produces a vertical beam, the most convenient choice for the angle of deflection

  18. Prospective Randomized Trial for Image-Guided Biopsy Using Cone-Beam CT Navigation Compared with Conventional CT.

    Science.gov (United States)

    Abi-Jaoudeh, Nadine; Fisher, Teresa; Jacobus, John; Skopec, Marlene; Radaelli, Alessandro; Van Der Bom, Imramsjah Martijn; Wesley, Robert; Wood, Bradford J

    2016-09-01

    To compare cone-beam computed tomography (CT) navigation vs conventional CT image guidance during biopsies. Patients scheduled for image-guided biopsies were prospectively and randomly assigned to conventional CT guidance vs cone-beam CT navigation. Radiation dose, accuracy of final needle position, rate of histopathologic diagnosis, and number of needle repositions to reach the target (defined as pullback to adjust position) were compared. A total of 58 patients (mean age, 57 y; 62.1% men) were randomized: 29 patients underwent 33 biopsies with CT guidance and 29 patients with 33 lesions underwent biopsy with cone-beam CT navigation. The average body mass index (BMI) was similar between groups, at 28.8 kg/m(2) ± 6.55 (P = .18). There was no difference between groups in terms of patient and lesion characteristics (eg, size, depth). The average lesion size was 29.1 ± 12.7mm for CT group vs 32.1mm ±16.8mm for cone-beam CT group (P cone-beam CT group was 0.3 ± 0.5, compared with 1.9 ± 2.3 with conventional CT (P cone-beam CT than with conventional CT (P cone-beam CT vs conventional CT (P = .018). Accuracy, defined as the difference between planned and final needle positions, was 4.9 mm ± 4.1 for the cone-beam CT group, compared with 12.2 mm ± 8.1 for conventional CT (P cone-beam CT (P = .67). Cone-beam CT navigation for biopsies improved targeting accuracy with fewer needle repositions, lower skin entry dose, and lower effective dose for planning scan, and a comparable histopathologic diagnosis rate. Copyright © 2016 SIR. All rights reserved.

  19. A megavoltage scatter correction technique for cone-beam CT images acquired during VMAT delivery

    International Nuclear Information System (INIS)

    Boylan, C J; Marchant, T E; Rowbottom, C G; Stratford, J; Rodgers, J; Malik, J; Choudhury, A; Shrimali, R

    2012-01-01

    Kilovoltage cone-beam CT (kV CBCT) can be acquired during the delivery of volumetric modulated arc therapy (VMAT), in order to obtain an image of the patient during treatment. However, the quality of such CBCTs is degraded by megavoltage (MV) scatter from the treatment beam onto the imaging panel. The objective of this paper is to introduce a novel MV scatter correction method for simultaneous CBCT during VMAT, and to investigate its effectiveness when compared to other techniques. The correction requires the acquisition of a separate set of images taken during VMAT delivery, while the kV beam is off. These images—which contain only the MV scatter contribution on the imaging panel—are then used to correct the corresponding kV/MV projections. To test this method, CBCTs were taken of an image quality phantom during VMAT delivery and measurements of contrast to noise ratio were made. Additionally, the correction was applied to the datasets of three VMAT prostate patients, who also received simultaneous CBCTs. The clinical image quality was assessed using a validated scoring system, comparing standard CBCTs to the uncorrected simultaneous CBCTs and a variety of correction methods. Results show that the correction is able to recover some of the low and high-contrast signal to noise ratio lost due to MV scatter. From the patient study, the corrected CBCT scored significantly higher than the uncorrected images in terms of the ability to identify the boundary between the prostate and surrounding soft tissue. In summary, a simple MV scatter correction method has been developed and, using both phantom and patient data, is shown to improve the image quality of simultaneous CBCTs taken during VMAT delivery. (paper)

  20. Estimation of effective dose from limited cone beam X-ray CT examination

    Energy Technology Data Exchange (ETDEWEB)

    Iwai, Kazuo; Arai, Yoshinori; Hashimoto, Koji [Nihon Univ., Tokyo (Japan). School of Dentistry; Nishizawa, Kanae

    2000-12-01

    The limited cone beam X-ray CT (Ortho-CT) was developed on the basis of multi-functional panoramic apparatus, SCANORA (Soredex Co. Helsinki Finland). The imaging intensifier (I.I.) was built in this apparatus as a X-ray detection device instead of X-ray film. The signal provided from I.I. was converted from analog into digital by an analog-digital converter and image reconstitution was done as a three-directional image of the dimensions 3.8 cm of width, 3.0 cm height and 3.8 cm depth with the personal computer. The 3DX Multi image micro CT'' (3DX) was developed along similar lines by MORITA Co., Ltd. (Kyoto, JAPAN). In this study, the stochastic effect on organ and tissue caused by examinations using Ortho-CT and 3DX was measured. The effective dose was estimated according to the recommendation of ICRP60 and was compared with those of panoramic radiography and computed tomography. The irradiation conditions were as follows: 85 kV, 10 mA with the filtration of 3 mmAl and added 1 mmCu for Ortho-CT, and 80 kV, 2 mA and the filtration of 3.1 mmAL for 3DX. The measurement of organ and tissue dose was performed using an anthropomorphic Rando woman phantom (Alderson Research Laboratories Co., Stanfora, CN), as well as by using two different type of thermoluminescent dosimeter (TLD); Panasonic UD-170A (BeO) and UD-110S (CaSO{sub 4}: Tm). The UD-170A was for dose measurement of the inner useful X-ray beams, while the UD-110S was for outer beams. The measured organ and tissue were those recommended with ICRP60 (gonad, breast, bone marrow, lung, thyroid gland, esophagus, stomach, colon, liver, bladder, skin, brain, thymus, adrenal, kidney, spleen, pancrease, upper large intestine, uterus, eyes and major salivary gland). The imaging by Orhto-CT was made in the left maxillary 1st molar, left mandibular 1st molar and temporomandibular joint. 3DX measurement was made in the maxillary incisor region and middle ear regions other than the regions mentioned above. The skin

  1. Calibration free beam hardening correction for cardiac CT perfusion imaging

    Science.gov (United States)

    Levi, Jacob; Fahmi, Rachid; Eck, Brendan L.; Fares, Anas; Wu, Hao; Vembar, Mani; Dhanantwari, Amar; Bezerra, Hiram G.; Wilson, David L.

    2016-03-01

    Myocardial perfusion imaging using CT (MPI-CT) and coronary CTA have the potential to make CT an ideal noninvasive gate-keeper for invasive coronary angiography. However, beam hardening artifacts (BHA) prevent accurate blood flow calculation in MPI-CT. BH Correction (BHC) methods require either energy-sensitive CT, not widely available, or typically a calibration-based method. We developed a calibration-free, automatic BHC (ABHC) method suitable for MPI-CT. The algorithm works with any BHC method and iteratively determines model parameters using proposed BHA-specific cost function. In this work, we use the polynomial BHC extended to three materials. The image is segmented into soft tissue, bone, and iodine images, based on mean HU and temporal enhancement. Forward projections of bone and iodine images are obtained, and in each iteration polynomial correction is applied. Corrections are then back projected and combined to obtain the current iteration's BHC image. This process is iterated until cost is minimized. We evaluate the algorithm on simulated and physical phantom images and on preclinical MPI-CT data. The scans were obtained on a prototype spectral detector CT (SDCT) scanner (Philips Healthcare). Mono-energetic reconstructed images were used as the reference. In the simulated phantom, BH streak artifacts were reduced from 12+/-2HU to 1+/-1HU and cupping was reduced by 81%. Similarly, in physical phantom, BH streak artifacts were reduced from 48+/-6HU to 1+/-5HU and cupping was reduced by 86%. In preclinical MPI-CT images, BHA was reduced from 28+/-6 HU to less than 4+/-4HU at peak enhancement. Results suggest that the algorithm can be used to reduce BHA in conventional CT and improve MPI-CT accuracy.

  2. Geometry calibration method for a cone-beam CT system.

    Science.gov (United States)

    Yang, Hongkai; Kang, Kejun; Xing, Yuxiang

    2017-05-01

    The positioning accuracy of each component is important to ensure the image quality of cone-beam CT. However, accurate positioning is not easy and requires experience and time. The option is to calibrate the geometric parameters and then plug them into a reconstruction algorithm which is the preferred solution in practice. In this case, the image quality is determined by the accuracy and precision of the calibration method. This work describes a method to independently calibrate an imaging system in each pose (projection angle) for a cone-beam CT with a nonideal circular trajectory. The calibration method uses a phantom with 12 beads on 2 planes that are observed on the radiographic images. This pose-independent calibration method (PIC) can decorrelate the relationships among the geometric parameters so that the parameters can be estimated one-by-one. This simplifies the calibration process. Besides the pose-independent calibration method, this paper also describes an extended calibration method with additional constraints on the system geometry. Both methods are validated with numerical simulations and then experimentally on a practical system with a scanning object loosely supported by rotating wheels. The object rotates during the CT data acquisition. The angular and pose information of the CT system are not accurately known a priori in this case. The numerical simulations and the experiments both provide satisfactory results. The relative error of the calibrated source-to-detector distance in the simulation is less than 0.1%. The errors in the calibrated roll, pitch, and yaw angles are less than 0.04°. A sensitivity study using various bead position uncertainties in random directions shows that the pose-independent calibration method is robust to measurement errors. Tests were also done with a nonideal circular trajectory for further validation. Images reconstructed using the geometric parameters from both the pose-independent and the extended calibration

  3. Cone beam CT for organs motion evaluation in pediatric abdominal neuroblastoma

    International Nuclear Information System (INIS)

    Nazmy, Mohamed Soliman; Khafaga, Yasser; Mousa, Amr; Khalil, Ehab

    2012-01-01

    Background and purpose: To quantify the organ motion relative to bone in different breathing states in pediatric neuroblastoma using cone beam CT (CBCT) for better definition of the planning margins during abdominal IMRT. Methods and materials: Forty-two datasets of kV CBCT for 9 pediatric patients with abdominal neuroblastoma treated with IMRT were evaluated. Organs positions on planning CT scan were considered the reference position against which organs and target motions were evaluated. The position of the kidneys and the liver was assessed in all scans. The target movement was evaluated in four patients who were treated for gross residual disease. Results: The mean age of the patients was 4.1 ± 1.6 years. The range of target movement in the craniocaudal direction (CC) was 5 mm. In the CC direction, the range of movement was 10 mm for the right kidney, and 8 mm for the left kidney. Similarly, the liver upper edge range of motion was 11 mm while the lower edge range of motion was 13 mm. Conclusions: With the use of daily CBCT we may be able to reduce the PTV margin. If CBCT is not used daily, a wider margin is needed.

  4. A robotic C-arm cone beam CT system for image-guided proton therapy: design and performance.

    Science.gov (United States)

    Hua, Chiaho; Yao, Weiguang; Kidani, Takao; Tomida, Kazuo; Ozawa, Saori; Nishimura, Takenori; Fujisawa, Tatsuya; Shinagawa, Ryousuke; Merchant, Thomas E

    2017-11-01

    A ceiling-mounted robotic C-arm cone beam CT (CBCT) system was developed for use with a 190° proton gantry system and a 6-degree-of-freedom robotic patient positioner. We report on the mechanical design, system accuracy, image quality, image guidance accuracy, imaging dose, workflow, safety and collision-avoidance. The robotic CBCT system couples a rotating C-ring to the C-arm concentrically with a kV X-ray tube and a flat-panel imager mounted to the C-ring. CBCT images are acquired with flex correction and maximally 360° rotation for a 53 cm field of view. The system was designed for clinical use with three imaging locations. Anthropomorphic phantoms were imaged to evaluate the image guidance accuracy. The position accuracy and repeatability of the robotic C-arm was high (beam proton systems with the added advantage of acquiring images at the treatment isocentre.

  5. Dose calculation based on Cone Beam CT images

    DEFF Research Database (Denmark)

    Slot Thing, Rune

    Cone beam CT (CBCT) imaging is frequently used in modern radiotherapy to ensure the proper positioning of the patient prior to each treatment fraction. With the increasing use of CBCT imaging for image guidance, interest has grown in exploring the potential use of these 3– or 4–D medical images...... image lag, scatter within the CBCT detector assembly, x-ray beam hardening from the patient, and truncation of the CBCT field of view were implemented for clinical CBCT imaging of lung cancer patients. Through the artefact corrections, Hounsfield Units in the CBCT images were recovered and shown...

  6. Full Scale Earth Fault Experiments on 10 kV laboratory network with comparative Measurements on Conventional CT's and VT's

    DEFF Research Database (Denmark)

    Sørensen, Stefan; Nielsen, Hans Ove; Bak-Jensen, Birgitte

    2002-01-01

    In this paper we present a result of a full scale earth fault carried out on the 10 kV research/laboratory distribution network at Kyndbyvaerket Denmark in May 2001. The network is compensated through a Petersen-Coil and current and voltage measurements were measured on conventional current...

  7. Low-dose megavoltage cone-beam CT for radiation therapy

    International Nuclear Information System (INIS)

    Pouliot, Jean; Bani-Hashemi, Ali; Chen, Josephine; Svatos, Michelle; Ghelmansarai, Farhad; Mitschke, Matthias; Aubin, Michele; Xia Ping; Morin, Olivier; Bucci, Kara; Roach, Mack; Hernandez, Paco; Zheng Zirao; Hristov, Dimitre; Verhey, Lynn

    2005-01-01

    Purpose: The objective of this work was to demonstrate the feasibility of acquiring low-exposure megavoltage cone-beam CT (MV CBCT) three-dimensional (3D) image data of sufficient quality to register the CBCT images to kilovoltage planning CT images for patient alignment and dose verification purposes. Methods and materials: A standard clinical 6-MV Primus linear accelerator, operating in arc therapy mode, and an amorphous-silicon (a-Si) flat-panel electronic portal-imaging device (EPID) were employed. The dose-pulse rate of a 6-MV Primus accelerator beam was windowed to expose an a-Si flat panel by using only 0.02 to 0.08 monitor units (MUs) per image. A triggered image-acquisition mode was designed to produce a high signal-to-noise ratio without pulsing artifacts. Several data sets were acquired for an anthropomorphic head phantom and frozen sheep and pig cadaver heads, as well as for a head-and-neck cancer patient on intensity-modulated radiotherapy (IMRT). For each CBCT image, a set of 90 to 180 projection images incremented by 1 deg to 2 deg was acquired. The two-dimensional (2D) projection images were then synthesized into a 3D image by use of cone-beam CT reconstruction. The resulting MV CBCT image set was used to visualize the 3D bony anatomy and some soft-tissue details. The 3D image registration with the kV planning CT was performed either automatically by application of a maximization of mutual information (MMI) algorithm or manually by aligning multiple 2D slices. Results: Low-noise 3D MV CBCT images without pulsing artifacts were acquired with a total delivered dose that ranged from 5 to 15 cGy. Acquisition times, including image readout, were on the order of 90 seconds for 180 projection images taken through a continuous gantry rotation of 180 deg . The processing time of the data required an additional 90 seconds for the reconstruction of a 256 3 cube with 1.0-mm voxel size. Implanted gold markers (1 mm x 3 mm) were easily visible for all exposure

  8. Triple-rule-out dual-source CT angiography of patients with acute chest pain: Dose reduction potential of 100 kV scanning

    International Nuclear Information System (INIS)

    Krissak, Radko; Henzler, Thomas; Prechel, Anne; Reichert, Miriam; Gruettner, Joachim; Sueselbeck, Tim; Schoenberg, Stefan O.; Fink, Christian

    2012-01-01

    Purpose: To investigate the dose reduction potential of low kV triple-rule-out dual-source CT angiography (TRO-CTA) in non-obese (BMI ≤ 25 kg/m 2 ) patients with acute chest pain. Materials and methods: Sixty consecutive patients were randomly assigned to two different retrospectively ECG-gated TRO-CTA protocols in this prospective trial: Thirty patients were examined with a 120-kV standard protocol (320 reference mAs with automatic tube current modulation, automatically adapted pitch and ECG-pulsing) and served as the control group (group 1), an otherwise identical 100 kV protocol was used in the other thirty patients (group 2) for a radiation dose reduction. Subjective image quality was assessed on a 5 point scale (1: excellent, 5: non-diagnostic) by two blinded observers. Quantitative image analysis assessed vascular attenuation, signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in different vascular segments. The effective dose was calculated from the dose length product (DLP) using a conversion coefficient of 0.017 mSv mGy −1 cm −1 . Results: There was no significant difference of age, BMI, heart rate, pitch or scan length between both patient groups. Subjective image quality was rated similar in both groups (group 1: 1.2 ± 0.4, group 2: average score = 1.3 ± 0.5). Vessel attenuation was significantly higher in group 2 than in group 1 (ascending aorta: 456 ± 83 HU vs. 370 ± 78 HU, p < 0.001; pulmonary artery: 468 ± 118 HU vs. 411 ± 91 HU, p = 0.03; left coronary artery: 437 ± 110 HU vs. 348 ± 89 HU, p < 0.001), however, there was no significant difference in SNR (13.2 ± 7.6 vs. 14.5 ± 7.5, p = 0.49) or CNR (13.8 ± 6.6 vs. 15.9 ± 7.7, p = 0.25). The effective radiation dose of the 100 kV protocol was significantly lower (9.6 ± 3.2 mSv vs. 18.1 ± 9.4 mSv, p < 0.0001). Conclusion: TRO-CTA with 100 kV is feasible in non-obese patients and results in diagnostic image quality and significantly reduced radiation dose.

  9. Cone-Beam CT with Fluoroscopic Overlay Versus Conventional CT Guidance for Percutaneous Abdominopelvic Abscess Drain Placement.

    Science.gov (United States)

    McKay, Tyler; Ingraham, Christopher R; Johnson, Guy E; Kogut, Matthew J; Vaidya, Sandeep; Padia, Siddharth A

    2016-01-01

    To compare technical success and procedure time for percutaneous abscess drain placement with fluoroscopic cone-beam computed tomography (CT) and two-axis needle guidance versus conventional CT guidance. A total of 85 consecutive patients undergoing abdominopelvic abscess drain placement guided by fluoroscopic cone-beam CT or conventional CT were retrospectively reviewed over a 2-year period. Forty-three patients underwent drain placement with cone-beam CT using XperGuide navigation and 42 underwent placement with conventional 64-slice CT. Patient characteristics, median abscess size (6.8 cm vs 7.8 cm; P = .14), and depth to abscess (7.2 cm vs 7.7 cm; P = .88) were similar between groups. Technical success rates were 98% (42 of 43) in the cone-beam CT group and 100% (42 of 42) in the conventional CT group (P = .32), with a 10-F pigtail drainage catheter inserted in the majority of cases. There were no complications in either group. There was no significant difference in effective dose between groups (9.6 mSv vs 10.7 mSv; P = .30). Procedure times were significantly shorter in the cone-beam CT group (43 min vs 62 min; P = .02). In addition, during the study period, there was a gradual improvement in procedure time in the cone-beam CT group (50% reduction), whereas procedure time did not change for the conventional CT group. Cone-beam CT guidance appears to be equivalent to conventional CT guidance for drain placement into medium-sized abdominopelvic collections, yielding similar technical success rates and radiation doses, with the additional benefit of reduced procedure times. Copyright © 2016 SIR. All rights reserved.

  10. Fast shading correction for cone beam CT in radiation therapy via sparse sampling on planning CT.

    Science.gov (United States)

    Shi, Linxi; Tsui, Tiffany; Wei, Jikun; Zhu, Lei

    2017-05-01

    The image quality of cone beam computed tomography (CBCT) is limited by severe shading artifacts, hindering its quantitative applications in radiation therapy. In this work, we propose an image-domain shading correction method using planning CT (pCT) as prior information which is highly adaptive to clinical environment. We propose to perform shading correction via sparse sampling on pCT. The method starts with a coarse mapping between the first-pass CBCT images obtained from the Varian TrueBeam system and the pCT. The scatter correction method embedded in the Varian commercial software removes some image errors but the CBCT images still contain severe shading artifacts. The difference images between the mapped pCT and the CBCT are considered as shading errors, but only sparse shading samples are selected for correction using empirical constraints to avoid carrying over false information from pCT. A Fourier-Transform-based technique, referred to as local filtration, is proposed to efficiently process the sparse data for effective shading correction. The performance of the proposed method is evaluated on one anthropomorphic pelvis phantom and 17 patients, who were scheduled for radiation therapy. (The codes of the proposed method and sample data can be downloaded from https://sites.google.com/view/linxicbct) RESULTS: The proposed shading correction substantially improves the CBCT image quality on both the phantom and the patients to a level close to that of the pCT images. On the phantom, the spatial nonuniformity (SNU) difference between CBCT and pCT is reduced from 74 to 1 HU. The root of mean square difference of SNU between CBCT and pCT is reduced from 83 to 10 HU on the pelvis patients, and from 101 to 12 HU on the thorax patients. The robustness of the proposed shading correction is fully investigated with simulated registration errors between CBCT and pCT on the phantom and mis-registration on patients. The sparse sampling scheme of our method successfully

  11. Cone beam CT paranasal sinuses versus standard multidetector and low dose multidetector CT studies.

    Science.gov (United States)

    Al Abduwani, Janan; ZilinSkiene, Laura; Colley, Steve; Ahmed, Shahzada

    2016-01-01

    Cone-beam computed tomography (CBCT) is a promising modality for quick outpatient imaging with lower radiation dose and less metal artifact when compared to conventional CT (MDCT) scans. This article will be providing evidence on the diagnostic and treatment-planning applications of CBCT in sinus imaging, mainly, in patients with chronic sinusitis for surgical planning, which retrospectively assessed 21 patients over a period of one year at Queen Elisabeth Hospital, Birmingham, UK. The main objective was to compare the absorbed dose of radiation from CBCT and conventional CT, and to compare the clarity and image quality for important structures in sinus anatomy in patients with sinus disease meriting CT scan imaging. Results of the mean effective dose of twenty-one consecutive CBCTs of paranasal sinuses were 0.27 mSv (range 0.05-0.48 mSv). This dose was approximately 40% lower when compared to a similar cohort of standard MDCT examinations and 30% lower when compared to low dose sinus CT scans. The visualization of high-contrast bone morphology on CBCT was comparable to standard sinus CT, allowing clear delineation of the principal surgically relevant osseous structures. Soft tissue visibility was however limited. We concluded that, CBCT scan provides a fast and efficient alternative to conventional CT with substantial radiation dose reduction and low dose MDCT techniques. However for more advanced sinus disease, conventional CT scan is preferable. The main objective was to compare the absorbed dose in the cone beam computed tomography (CBCT) and conventional CT and to compare the clarity and image quality for important structures in sinus anatomy. Retrospective scan review. We compared both the image characteristics and the effective dose used in CBCT and conventional MDCT sinus imaging, performed at Queen Elizabeth Hospital Birmingham over a 1 year period. Patient with sinus disease meriting CT scan imaging. Comparison of effective dose exposure and image quality

  12. Fast 3D multiple fan-beam CT systems

    Science.gov (United States)

    Kohlbrenner, Adrian; Haemmerle, Stefan; Laib, Andres; Koller, Bruno; Ruegsegger, Peter

    1999-09-01

    Two fast, CCD-based three-dimensional CT scanners for in vivo applications have been developed. One is designed for small laboratory animals and has a voxel size of 20 micrometer, while the other, having a voxel size of 80 micrometer, is used for human examinations. Both instruments make use of a novel multiple fan-beam technique: radiation from a line-focus X-ray tube is divided into a stack of fan-beams by a 28 micrometer pitch foil collimator. The resulting wedge-shaped X-ray field is the key to the instrument's high scanning speed and allows to position the sample close to the X-ray source, which makes it possible to build compact CT systems. In contrast to cone- beam scanners, the multiple fan-beam scanner relies on standard fan-beam algorithms, thereby eliminating inaccuracies in the reconstruction process. The projections from one single rotation are acquired within 2 min and are subsequently reconstructed into a 1024 X 1024 X 255 voxel array. Hence a single rotation about the sample delivers a 3D image containing a quarter of a billion voxels. Such volumetric images are 6.6 mm in height and can be stacked on top of each other. An area CCD sensor bonded to a fiber-optic light guide acts as a detector. Since no image intensifier, conventional optics or tapers are used throughout the system, the image is virtually distortion free. The scanner's high scanning speed and high resolution at moderately low radiation dose are the basis for reliable time serial measurements and analyses.

  13. Segmentation-free empirical beam hardening correction for CT

    Energy Technology Data Exchange (ETDEWEB)

    Schüller, Sören; Sawall, Stefan [German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, Heidelberg 69120 (Germany); Stannigel, Kai; Hülsbusch, Markus; Ulrici, Johannes; Hell, Erich [Sirona Dental Systems GmbH, Fabrikstraße 31, 64625 Bensheim (Germany); Kachelrieß, Marc, E-mail: marc.kachelriess@dkfz.de [German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg (Germany)

    2015-02-15

    Purpose: The polychromatic nature of the x-ray beams and their effects on the reconstructed image are often disregarded during standard image reconstruction. This leads to cupping and beam hardening artifacts inside the reconstructed volume. To correct for a general cupping, methods like water precorrection exist. They correct the hardening of the spectrum during the penetration of the measured object only for the major tissue class. In contrast, more complex artifacts like streaks between dense objects need other techniques of correction. If using only the information of one single energy scan, there are two types of corrections. The first one is a physical approach. Thereby, artifacts can be reproduced and corrected within the original reconstruction by using assumptions in a polychromatic forward projector. These assumptions could be the used spectrum, the detector response, the physical attenuation and scatter properties of the intersected materials. A second method is an empirical approach, which does not rely on much prior knowledge. This so-called empirical beam hardening correction (EBHC) and the previously mentioned physical-based technique are both relying on a segmentation of the present tissues inside the patient. The difficulty thereby is that beam hardening by itself, scatter, and other effects, which diminish the image quality also disturb the correct tissue classification and thereby reduce the accuracy of the two known classes of correction techniques. The herein proposed method works similar to the empirical beam hardening correction but does not require a tissue segmentation and therefore shows improvements on image data, which are highly degraded by noise and artifacts. Furthermore, the new algorithm is designed in a way that no additional calibration or parameter fitting is needed. Methods: To overcome the segmentation of tissues, the authors propose a histogram deformation of their primary reconstructed CT image. This step is essential for the

  14. Segmental and subsegmental pulmonary arteries: evaluation with electron-beam versus spiral CT.

    Science.gov (United States)

    Schoepf, U J; Helmberger, T; Holzknecht, N; Kang, D S; Bruening, R D; Aydemir, S; Becker, C R; Muehling, O; Knez, A; Haberl, R; Reiser, M F

    2000-02-01

    To compare contrast agent-enhanced spiral and electron-beam computed tomography (CT) for the analysis of segmental and subsegmental pulmonary arteries. CT angiography of the pulmonary arteries was performed in 56 patients to rule out pulmonary embolism. Electron-beam CT was performed in 28 patients. The other 28 patients underwent spiral CT with comparable scanning protocols. The depiction of segmental and subsegmental arteries was analyzed by three independent readers. The contrast enhancement in the main pulmonary artery was measured in each patient. Analysis was performed in 1,120 segmental and 2, 240 subsegmental arteries. One segmental (RA7, P =.010) and two subsegmental (LA7b, P =.029; RA6a+b, P =.038) arteries in paracardiac and basal segments of the lung were depicted significantly better with electron-beam CT. There was no statistically significant difference between electron-beam and spiral CT in the total number of analyzable peripheral arteries depicted. The mean contrast enhancement in the main pulmonary artery was 362 HU in electron-beam CT studies versus 248 HU in spiral CT studies. Detailed visualization of peripheral pulmonary arteries is well within the scope of advanced CT techniques. Electron-beam CT has minor advantages in analyzing paracardiac arteries, probably because of reduction of motion artifacts and higher contrast enhancement. Further studies are needed to establish whether electron-beam CT allows a more confident diagnosis of emboli in these vessels.

  15. Beam hardening correction in CT myocardial perfusion measurement

    Science.gov (United States)

    So, Aaron; Hsieh, Jiang; Li, Jian-Ying; Lee, Ting-Yim

    2009-05-01

    This paper presents a method for correcting beam hardening (BH) in cardiac CT perfusion imaging. The proposed algorithm works with reconstructed images instead of projection data. It applies thresholds to separate low (soft tissue) and high (bone and contrast) attenuating material in a CT image. The BH error in each projection is estimated by a polynomial function of the forward projection of the segmented image. The error image is reconstructed by back-projection of the estimated errors. A BH-corrected image is then obtained by subtracting a scaled error image from the original image. Phantoms were designed to simulate the BH artifacts encountered in cardiac CT perfusion studies of humans and animals that are most commonly used in cardiac research. These phantoms were used to investigate whether BH artifacts can be reduced with our approach and to determine the optimal settings, which depend upon the anatomy of the scanned subject, of the correction algorithm for patient and animal studies. The correction algorithm was also applied to correct BH in a clinical study to further demonstrate the effectiveness of our technique.

  16. Pediatric Percutaneous Osteoid Osteoma Ablation: Cone-Beam CT with Fluoroscopic Overlay Versus Conventional CT Guidance.

    Science.gov (United States)

    Perry, Brandon C; Monroe, Eric J; McKay, Tyler; Kanal, Kalpana M; Shivaram, Giridhar

    2017-10-01

    To compare technical success, clinical success, complications, radiation dose, and total room utilization time for osteoid osteoma thermal (radiofrequency or microwave) ablation using cone-beam computed tomography (CBCT) with two-axis fluoroscopic navigational overlay versus conventional computed tomography (CT) guidance. A retrospective review was performed to identify all osteoid osteoma ablations performed over a 5.5-year period at a single tertiary care pediatric hospital. Twenty-five ablations (15 radiofrequency and 10 microwave) in 23 patients undergoing fluoroscopic CBCT-guided osteoid osteoma ablation were compared to 35 ablations (35 radiofrequency) in 32 patients undergoing ablation via conventional CT guidance. Dose area product and dose length product were recorded for CBCT and conventional CT, respectively, and converted to effective doses. Technical success, clinical success (cessation of pain and medication use 1 month after ablation), complications, radiation dose, and total room utilization time were compared. All procedures were technically successful. Twenty-two of 25 (88.0%) CBCT and 31 of 35 (88.6%) conventional CT-guided ablations achieved immediate clinical success. There were two minor complications in each group and no major complications. Mean effective radiation dose was significantly lower for CBCT compared to CT guidance (0.12 vs. 0.39 mSv, p = 0.02). Mean total room utilization time for CBCT was longer (133.5 vs. 97.5 min, p = 0.0001). Fluoroscopic CBCT guidance for percutaneous osteoid osteoma ablation yields similar technical and clinical success, reduced radiation dose, and increased total room utilization time compared to conventional CT guidance.

  17. Helical cone beam CT with an asymmetrical detector

    International Nuclear Information System (INIS)

    Zamyatin, Alexander A.; Taguchi, Katsuyuki; Silver, Michael D.

    2005-01-01

    If a multislice or other area detector is shifted to one side to cover a larger field of view, then the data are truncated on one side. We propose a method to restore the missing data in helical cone-beam acquisitions that uses measured data on the longer side of the asymmetric detector array. The method is based on the idea of complementary rays, which is well known in fan beam geometry; in this paper we extend this concept to the cone-beam case. Different cases of complementary data coverage and dependence on the helical pitch are considered. The proposed method is used in our prototype 16-row CT scanner with an asymmetric detector and a 700 mm field of view. For evaluation we used scanned body phantom data and computer-simulated data. To simulate asymmetric truncation, the full, symmetric datasets were truncated by dropping either 22.5% or 45% from one side of the detector. Reconstructed images from the prototype scanner with the asymmetrical detector show excellent image quality in the extended field of view. The proposed method allows flexible helical pitch selection and can be used with overscan, short-scan, and super-short-scan reconstructions

  18. A kernel-based dose calculation algorithm for kV photon beams with explicit handling of energy and material dependencies.

    Science.gov (United States)

    Reinhart, Anna Merle; Fast, Martin F; Ziegenhein, Peter; Nill, Simeon; Oelfke, Uwe

    2017-01-01

    Mimicking state-of-the-art patient radiotherapy with high-precision irradiators for small animals is expected to advance the understanding of dose-effect relationships and radiobiology in general. We work on the implementation of intensity-modulated radiotherapy-like irradiation schemes for small animals. As a first step, we present a fast analytical dose calculation algorithm for keV photon beams. We follow a superposition-convolution approach adapted to kV X-rays, based on previous work for microbeam therapy. We assume local energy deposition at the photon interaction point due to the short electron ranges in tissue. This allows us to separate the dose calculation into locally absorbed primary dose and the scatter contribution, calculated in a point kernel approach. We validate our dose model against Geant4 Monte Carlo (MC) simulations and compare the results to Muriplan (XStrahl Ltd, Camberley, UK). For field sizes of (1 mm) 2 to (1 cm) 2 in water, the depth dose curves show a mean disagreement of 1.7% to MC simulations, with the largest deviations in the entrance region (4%) and at large depths (5% at 7 cm). Larger discrepancies are observed at water-to-bone boundaries, in bone and at the beam edges in slab phantoms and a mouse brain. Calculation times are in the order of 5 s for a single beam. The algorithm shows good agreement with MC simulations in an initial validation. It has the potential to become an alternative to full MC dose calculation. Advances in knowledge: The presented algorithm demonstrates the potential of kernel-based dose calculation for kV photon beams. It will be valuable in intensity-modulated radiotherapy and inverse treatment planning for high precision small-animal radiotherapy.

  19. Shading correction assisted iterative cone-beam CT reconstruction

    Science.gov (United States)

    Yang, Chunlin; Wu, Pengwei; Gong, Shutao; Wang, Jing; Lyu, Qihui; Tang, Xiangyang; Niu, Tianye

    2017-11-01

    Recent advances in total variation (TV) technology enable accurate CT image reconstruction from highly under-sampled and noisy projection data. The standard iterative reconstruction algorithms, which work well in conventional CT imaging, fail to perform as expected in cone beam CT (CBCT) applications, wherein the non-ideal physics issues, including scatter and beam hardening, are more severe. These physics issues result in large areas of shading artifacts and cause deterioration to the piecewise constant property assumed in reconstructed images. To overcome this obstacle, we incorporate a shading correction scheme into low-dose CBCT reconstruction and propose a clinically acceptable and stable three-dimensional iterative reconstruction method that is referred to as the shading correction assisted iterative reconstruction. In the proposed method, we modify the TV regularization term by adding a shading compensation image to the reconstructed image to compensate for the shading artifacts while leaving the data fidelity term intact. This compensation image is generated empirically, using image segmentation and low-pass filtering, and updated in the iterative process whenever necessary. When the compensation image is determined, the objective function is minimized using the fast iterative shrinkage-thresholding algorithm accelerated on a graphic processing unit. The proposed method is evaluated using CBCT projection data of the Catphan© 600 phantom and two pelvis patients. Compared with the iterative reconstruction without shading correction, the proposed method reduces the overall CT number error from around 200 HU to be around 25 HU and increases the spatial uniformity by a factor of 20 percent, given the same number of sparsely sampled projections. A clinically acceptable and stable iterative reconstruction algorithm for CBCT is proposed in this paper. Differing from the existing algorithms, this algorithm incorporates a shading correction scheme into the low

  20. Cone Beam Optical CT Investigation on Tissue Equivalent Normoxic Polymer Gel Dosimeter

    International Nuclear Information System (INIS)

    Kumar, D Senthil; Samuel, E James Jebaseelan

    2010-01-01

    A potential method has emerged in the form of water-equivalent '3D gel dosimetry' using optical computed-tomography (optical-CT) which enables accurate, high resolution, 3D measurement of dose distributions associated with modern radiation treatments.. Optical Cone Beam CT (CBCT) scanner plays a major role for Gel dosimeter readout and clinical radiation therapy as 3-Dimensional Radiation Dosimetry. The normoxic PAGAT (Polyacrylamide Gelatin and Tetrakis) gel is used as a dosimeter for this cone beam CT analysis due to its tissue equivalent behaviour. Applying a uniform background subtraction of open field intensity resulted in cone beam CT reconstructed attenuation coefficient for a PAGAT Gel Dosimeter.

  1. Computer aided breast density evaluation in cone beam breast CT

    Science.gov (United States)

    Zhang, Xiaohua; Ning, Ruola

    2011-03-01

    Cone Beam Breast CT is a three-dimensional breast imaging modality with high contrast resolution and no tissue overlap. With these advantages, it is possible to measure volumetric breast density accurately and quantitatively with CBBCT 3D images. Three major breast components need to be segmented: skin, fat and glandular tissue. In this research, a modified morphological processing is applied to the CBBCT images to detect and remove the skin of the breast. After the skin is removed, a 2-step fuzzy clustering scheme is applied to the CBBCT image volume to adaptively cluster the image voxels into fat and glandular tissue areas based on the intensity of each voxel. Finally, the CBBCT breast volume images are divided into three categories: skin, fat and glands. Clinical data is used and the quantitative CBBCT breast density evaluation results are compared with the mammogram-based BIRADS breast density categories.

  2. Electron-beam CT diagnosis of congenital cardiovascular diverticula

    International Nuclear Information System (INIS)

    Yang Youyou; Zheng Lili; Li Xiangmin; Zhou Xuhui; Peng Qian; Meng Quanfei; Dai Ruping

    2008-01-01

    Objective: To investigate the clinical application of electron-beam CT (EBCT) in the diagnosis of congenital cardiovascular diverticula. Methods: Retrospective analysis of 9 patients with congenital cardiovascular diverticula confirmed by operation and pathology was done. Of them, enhanced continuous volume scan was performed on 8 patients and enhanced single slice scan was performed on one patient with an Imatron C-150 scanner. Results: The group of 9 patients included one patient with diverticulum of the left ventricle, 3 patients with diverticulum of the atria and 5 patients with diverticulum of the aorta. EBCT scan and three dimensional reconstruction could demonstrate not only the origin, size, shape, location and adjacent structure of diverticula, but also other important complicated abnormalities such as ventriculoarterial connection disorder, cardiac septal defect, aortic coarctation and even dissection. Conclusion: EBCT is an ideal noninvasive technique in the diagnosis of congenital cardiovascular diverticula. (authors)

  3. 3D dictionary learning based iterative cone beam CT reconstruction

    Directory of Open Access Journals (Sweden)

    Ti Bai

    2014-03-01

    Full Text Available Purpose: This work is to develop a 3D dictionary learning based cone beam CT (CBCT reconstruction algorithm on graphic processing units (GPU to improve the quality of sparse-view CBCT reconstruction with high efficiency. Methods: A 3D dictionary containing 256 small volumes (atoms of 3 × 3 × 3 was trained from a large number of blocks extracted from a high quality volume image. On the basis, we utilized cholesky decomposition based orthogonal matching pursuit algorithm to find the sparse representation of each block. To accelerate the time-consuming sparse coding in the 3D case, we implemented the sparse coding in a parallel fashion by taking advantage of the tremendous computational power of GPU. Conjugate gradient least square algorithm was adopted to minimize the data fidelity term. Evaluations are performed based on a head-neck patient case. FDK reconstruction with full dataset of 364 projections is used as the reference. We compared the proposed 3D dictionary learning based method with tight frame (TF by performing reconstructions on a subset data of 121 projections. Results: Compared to TF based CBCT reconstruction that shows good overall performance, our experiments indicated that 3D dictionary learning based CBCT reconstruction is able to recover finer structures, remove more streaking artifacts and also induce less blocky artifacts. Conclusion: 3D dictionary learning based CBCT reconstruction algorithm is able to sense the structural information while suppress the noise, and hence to achieve high quality reconstruction under the case of sparse view. The GPU realization of the whole algorithm offers a significant efficiency enhancement, making this algorithm more feasible for potential clinical application.-------------------------------Cite this article as: Bai T, Yan H, Shi F, Jia X, Lou Y, Xu Q, Jiang S, Mou X. 3D dictionary learning based iterative cone beam CT reconstruction. Int J Cancer Ther Oncol 2014; 2(2:020240. DOI: 10

  4. Adaptive radiotherapy based on contrast enhanced cone beam CT imaging

    Energy Technology Data Exchange (ETDEWEB)

    Soevik, Aaste; Skogmo, Hege K. (Dept. of Companion Animal Clinical Sciences, Norwegian School of Veterinary Science, Oslo (Norway)), E-mail: aste.sovik@nvh.no; Roedal, Jan (Dept. of Companion Animal Clinical Sciences, Norwegian School of Veterinary Science, Oslo (Norway)); Lervaag, Christoffer; Eilertsen, Karsten; Malinen, Eirik (Dept. of Medical Physics, The Norwegian Radium Hospital, Oslo Univ. Hospital, Oslo (Norway))

    2010-10-15

    Cone beam CT (CBCT) imaging has become an integral part of radiation therapy, with images typically used for offline or online patient setup corrections based on bony anatomy co-registration. Ideally, the co-registration should be based on tumor localization. However, soft tissue contrast in CBCT images may be limited. In the present work, contrast enhanced CBCT (CECBCT) images were used for tumor visualization and treatment adaptation. Material and methods. A spontaneous canine maxillary tumor was subjected to repeated cone beam CT imaging during fractionated radiotherapy (10 fractions in total). At five of the treatment fractions, CECBCT images, employing an iodinated contrast agent, were acquired, as well as pre-contrast CBCT images. The tumor was clearly visible in post-contrast minus pre-contrast subtraction images, and these contrast images were used to delineate gross tumor volumes. IMRT dose plans were subsequently generated. Four different strategies were explored: 1) fully adapted planning based on each CECBCT image series, 2) planning based on images acquired at the first treatment fraction and patient repositioning following bony anatomy co-registration, 3) as for 2), but with patient repositioning based on co-registering contrast images, and 4) a strategy with no patient repositioning or treatment adaptation. The equivalent uniform dose (EUD) and tumor control probability (TCP) calculations to estimate treatment outcome for each strategy. Results. Similar translation vectors were found when bony anatomy and contrast enhancement co-registration were compared. Strategy 1 gave EUDs closest to the prescription dose and the highest TCP. Strategies 2 and 3 gave EUDs and TCPs close to that of strategy 1, with strategy 3 being slightly better than strategy 2. Even greater benefits from strategies 1 and 3 are expected with increasing tumor movement or deformation during treatment. The non-adaptive strategy 4 was clearly inferior to all three adaptive strategies

  5. Adaptive radiotherapy based on contrast enhanced cone beam CT imaging

    International Nuclear Information System (INIS)

    Soevik, Aaste; Skogmo, Hege K.; Roedal, Jan; Lervaag, Christoffer; Eilertsen, Karsten; Malinen, Eirik

    2010-01-01

    Cone beam CT (CBCT) imaging has become an integral part of radiation therapy, with images typically used for offline or online patient setup corrections based on bony anatomy co-registration. Ideally, the co-registration should be based on tumor localization. However, soft tissue contrast in CBCT images may be limited. In the present work, contrast enhanced CBCT (CECBCT) images were used for tumor visualization and treatment adaptation. Material and methods. A spontaneous canine maxillary tumor was subjected to repeated cone beam CT imaging during fractionated radiotherapy (10 fractions in total). At five of the treatment fractions, CECBCT images, employing an iodinated contrast agent, were acquired, as well as pre-contrast CBCT images. The tumor was clearly visible in post-contrast minus pre-contrast subtraction images, and these contrast images were used to delineate gross tumor volumes. IMRT dose plans were subsequently generated. Four different strategies were explored: 1) fully adapted planning based on each CECBCT image series, 2) planning based on images acquired at the first treatment fraction and patient repositioning following bony anatomy co-registration, 3) as for 2), but with patient repositioning based on co-registering contrast images, and 4) a strategy with no patient repositioning or treatment adaptation. The equivalent uniform dose (EUD) and tumor control probability (TCP) calculations to estimate treatment outcome for each strategy. Results. Similar translation vectors were found when bony anatomy and contrast enhancement co-registration were compared. Strategy 1 gave EUDs closest to the prescription dose and the highest TCP. Strategies 2 and 3 gave EUDs and TCPs close to that of strategy 1, with strategy 3 being slightly better than strategy 2. Even greater benefits from strategies 1 and 3 are expected with increasing tumor movement or deformation during treatment. The non-adaptive strategy 4 was clearly inferior to all three adaptive strategies

  6. Effective dose from cone beam CT examinations in dentistry.

    Science.gov (United States)

    Roberts, J A; Drage, N A; Davies, J; Thomas, D W

    2009-01-01

    Cone beam CT (CBCT) is becoming an increasingly utilized imaging modality for dental examinations in the UK. Previous studies have presented little information on patient dose for the range of fields of view (FOVs) that can be utilized. The purpose of the study was therefore to calculate the effective dose delivered to the patient during a selection of CBCT examinations performed in dentistry. In particular, the i-CAT CBCT scanner was investigated for several imaging protocols commonly used in clinical practice. A Rando phantom containing thermoluminescent dosemeters was scanned. Using both the 1990 and recently approved 2007 International Commission on Radiological Protection recommended tissue weighting factors, effective doses were calculated. The doses (E(1990), E(2007)) were: full FOV head (92.8 microSv, 206.2 microSv); 13 cm scan of the jaws (39.5 microSv, 133.9 microSv); 6 cm high-resolution mandible (47.2 microSv, 188.5 microSv); 6 cm high-resolution maxilla (18.5 microSv, 93.3 microSv); 6 cm standard mandible (23.9 microSv, 96.2 microSv); and 6 cm standard maxilla (9.7 microSv, 58.9 microSv). The doses from CBCT are low compared with conventional CT but significantly higher than conventional dental radiography techniques.

  7. Distribution and Detection of Radiopaque Beads after Hepatic Transarterial Embolization in Swine: Cone-Beam CT versus MicroCT.

    Science.gov (United States)

    Thompson, John G; van der Sterren, William; Bakhutashvili, Ivane; van der Bom, Imramsjah M; Radaelli, Alessandro G; Karanian, John W; Esparza-Trujillo, Juan; Woods, David L; Lewis, Andrew L; Wood, Bradford J; Pritchard, William F

    2018-04-01

    To determine the true distribution of radiopaque beads (ROBs) after hepatic embolization in swine as imaged by micro-computed tomography (microCT) compared with in vivo cone-beam computerized tomography (CT) imaged at different kVp settings. Swine (n = 3) underwent hepatic transarterial embolization (n = 6) with the use of 70-150-μm ROBs under fluoroscopic guidance. After stasis, in vivo cone-beam CT was performed at 120, 100, and 80 kVp. The animal was euthanized, the liver resected, and microCT with 17 μm resolution performed on embolized tissue samples. The resulting cone-beam CT and microCT data were segmented and registered. Total vessel length, minimum volume-enclosing ellipsoid (MVEE), and number of independent volumes were measured. Maximum-intensity projections (MIPs) were generated for each cone-beam CT. Metrics for all cone-beam CT segmentations differed significantly from microCT segmentations. Segmentations at 80 kVp presented significantly greater vessel length, MVEE, and number of independent volumes compared with 100 kVp and 120 kVp. In addition, 100 kVp segmentations presented significantly greater vessel length than 120 kVp. MIPs presented greater visualization than cone-beam CT segmentations and improved as kVp decreased. The full ROB distribution was more extensive than was apparent on cone-beam CT. Quantitative measures of embolic distribution demonstrated significantly better correlation with microCT with decreasing kVp. Similarly, qualitative analysis of MIPs showed improved visualization of beads with decreasing kVp. These findings demonstrate the clinical value of 80 kVp and 100 kVp protocols in the imaging of radiopaque embolizations compared with 120 kVp. However, considerations on X-ray penetration and dose may favor use of 100 kVp imaging over 80 kVp. Published by Elsevier Inc.

  8. SU-F-J-205: Effect of Cone Beam Factor On Cone Beam CT Number Accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Yao, W; Hua, C; Farr, J; Brady, S; Merchant, T [St. Jude Children’s Research Hospital, Memphis, TN (United States)

    2016-06-15

    Purpose: To examine the suitability of a Catphan™ 700 phantom for image quality QA of a cone beam computed tomography (CBCT) system deployed for proton therapy. Methods: Catphan phantoms, particularly Catphan™ 504, are commonly used in image quality QA for CBCT. As a newer product, Catphan™ 700 offers more tissue equivalent inserts which may be useful for generating the electron density – CT number curve for CBCT based treatment planning. The sensitometry-and-geometry module used in Catphan™ 700 is located at the end of the phantom and after the resolution line pair module. In Catphan™ 504 the line pair module is located at the end of the phantom and after the sensitometry-and-geometry module. To investigate the effect of difference in location on CT number accuracy due to the cone beam factor, we scanned the Catphan™ 700 with the central plane of CBCT at the center of the phantom, line pair and sensitometry-andgeometry modules of the phantom, respectively. The protocol head and thorax scan modes were used. For each position, scans were repeated 4 times. Results: For the head scan mode, the standard deviation (SD) of the CT numbers of each insert under 4 repeated scans was up to 20 HU, 11 HU, and 11 HU, respectively, for the central plane of CBCT located at the center of the phantom, line pair, and sensitometry-and-geometry modules of the phantom. The mean of the SD was 9.9 HU, 5.7 HU, and 5.9 HU, respectively. For the thorax mode, the mean of the SD was 4.5 HU, 4.4 HU, and 4.4 HU, respectively. The assessment of image quality based on resolution and spatial linearity was not affected by imaging location changes. Conclusion: When the Catphan™ 700 was aligned to the center of imaging region, the CT number accuracy test may not meet expectations. We recommend reconfiguration of the modules.

  9. Commissioning and operation of 130 kV/130 A switched-mode HV power supplies with the upgraded JET neutral beam injectors

    International Nuclear Information System (INIS)

    Edwards, D.C.; Bigi, M.; Brown, D.P.D.; Ganuza, D.; Garcia, F.; Hudson, Z.; Jones, T.T.C.; Perez, A.

    2005-01-01

    The design features, on-site testing, commissioning and operation are described of two new 130 kV/130 A HV power supply units serving four upgraded 130 kV/60 A positive ion neutral injectors (PINIs) on JET. Both units were factory tested at full power and pulse length into dummy resistive load. Following on-site installation, the factory tests were repeated. The transition from dummy-load testing to PINI operation required full integration of the HVPS within the overall JET control system, and rigorous testing of the co-ordinated actions and protections of all PINI power supplies (filament and arc for plasma source and negative suppression grid). The implementation of these functions is described. Extensive use was made of parasitic integrated test pulses, where the other PINIs could be operated normally, with the HVPS energised under full remote control together with the corresponding PINI plasma sources, but with the HVPS connected to dummy load. The amount of NB operation time dedicated to commissioning was thereby minimised, yet gave a high degree of confidence of readiness for HV energisation of the PINI, and first beam operation followed less than 24 h from HV connection to the PINI. The routine operating experience and performance, including load protection characteristics of the new HVPS units are also described

  10. Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scanner

    International Nuclear Information System (INIS)

    Geleijns, J; Bruin, P W de; Salvado Artells, M; Mather, R; Muramatsu, Y; McNitt-Gray, M F

    2009-01-01

    Computed tomography (CT) dosimetry should be adapted to the rapid developments in CT technology. Recently a 160 mm wide, 320 detector row, cone beam CT scanner that challenges the existing Computed Tomography Dose Index (CTDI) dosimetry paradigm was introduced. The purpose of this study was to assess dosimetric characteristics of this cone beam scanner, to study the appropriateness of existing CT dose metrics and to suggest a pragmatic approach for CT dosimetry for cone beam scanners. Dose measurements with a small Farmer-type ionization chamber and with 100 mm and 300 mm long pencil ionization chambers were performed free in air to characterize the cone beam. According to the most common dose metric in CT, namely CTDI, measurements were also performed in 150 mm and 350 mm long CT head and CT body dose phantoms with 100 mm and 300 mm long pencil ionization chambers, respectively. To explore effects that cannot be measured with ionization chambers, Monte Carlo (MC) simulations of the dose distribution in 150 mm, 350 mm and 700 mm long CT head and CT body phantoms were performed. To overcome inconsistencies in the definition of CTDI 100 for the 160 mm wide cone beam CT scanner, doses were also expressed as the average absorbed dose within the pencil chamber (D-bar 100 ). Measurements free in air revealed excellent correspondence between CTDI 300air and D-bar 100air , while CTDI 100air substantially underestimates CTDI 300air . Results of measurements in CT dose phantoms and corresponding MC simulations at centre and peripheral positions were weighted and revealed good agreement between CTDI 300w , D-bar 100w and CTDI 600w , while CTDI 100w substantially underestimates CTDI 300w . D-bar 100w provides a pragmatic metric for characterizing the dose of the 160 mm wide cone beam CT scanner. This quantity can be measured with the widely available 100 mm pencil ionization chamber within 150 mm long CT dose phantoms. CTDI 300w measured in 350 mm long CT dose phantoms serves

  11. Auto calibration of a cone-beam-CT

    International Nuclear Information System (INIS)

    Gross, Daniel; Heil, Ulrich; Schulze, Ralf; Schoemer, Elmar; Schwanecke, Ulrich

    2012-01-01

    Purpose: This paper introduces a novel autocalibration method for cone-beam-CTs (CBCT) or flat-panel CTs, assuming a perfect rotation. The method is based on ellipse-fitting. Autocalibration refers to accurate recovery of the geometric alignment of a CBCT device from projection images alone, without any manual measurements. Methods: The authors use test objects containing small arbitrarily positioned radio-opaque markers. No information regarding the relative positions of the markers is used. In practice, the authors use three to eight metal ball bearings (diameter of 1 mm), e.g., positioned roughly in a vertical line such that their projection image curves on the detector preferably form large ellipses over the circular orbit. From this ellipse-to-curve mapping and also from its inversion the authors derive an explicit formula. Nonlinear optimization based on this mapping enables them to determine the six relevant parameters of the system up to the device rotation angle, which is sufficient to define the geometry of a CBCT-machine assuming a perfect rotational movement. These parameters also include out-of-plane rotations. The authors evaluate their method by simulation based on data used in two similar approaches [L. Smekal, M. Kachelriess, S. E, and K. Wa, “Geometric misalignment and calibration in cone-beam tomography,” Med. Phys. 31(12), 3242–3266 (2004); K. Yang, A. L. C. Kwan, D. F. Miller, and J. M. Boone, “A geometric calibration method for cone beam CT systems,” Med. Phys. 33(6), 1695–1706 (2006)]. This allows a direct comparison of accuracy. Furthermore, the authors present real-world 3D reconstructions of a dry human spine segment and an electronic device. The reconstructions were computed from projections taken with a commercial dental CBCT device having two different focus-to-detector distances that were both calibrated with their method. The authors compare their reconstruction with a reconstruction computed by the manufacturer of the

  12. Fundamental efficiency of limited cone-beam X-ray CT (3DX Multi image micro CT) for practical use

    Energy Technology Data Exchange (ETDEWEB)

    Arai, Yoshinori; Hashimoto, Koji; Iwai, Kazuo; Shinoda, Koji [Nihon Univ., Tokyo (Japan). School of Dentistry

    2000-06-01

    The limited cone-beam CT known as Ortho-CT has been used in clinical examination since 1997. On the basis of this experience, we developed a new style of limited cone-beam CT for practical use, called ''3DX Multi image micro CT'' (3DX). The purpose of this study was to introduce this new style of limited cone-beam X-CT by comparing it to the prototype (Ortho-CT). 3DX was compared with Ortho-CT regarding the fundamental efficiency. The skin doses of 3DX and Ortho-CT were measured using TLD on a Rando phantom. The resolutions of both systems were evaluated with MTF (modulation transfer function). The subjective image quality was evaluated on the following anatomical landmarks: the inner ear, temporomandibular joint (TMJ), maxillary first molar, mandibular first molar. Five dental radiologists and two otolaryngologists evaluated the quality of 3DX images comparing then with that of Ortho-CT images for the same observation point. The five-point scale ranged from one point (inferior) to five (superior). The skin doses were 1.07 mSv with 3DX and 1.19 mSv with Ortho-CT. The skin dose of ''3DX'' was almost the same as that of Ortho-CT. The resolutions of Ortho-CT were 0.6 line pair/mm (horizontal) and 0.9 line pair/mm (vertical) on the MTF (0.5). The resolution of 3DX was 1.1 line pair/mm (horizontal) and 1.3 line pair/mm (vertical). The subjective image quality of 3DX was better than that of Ortho-CT at every observation point. The minimum score was 3.46, and maximum score was 4.17. There were significant differences with every observation point (p<0.05). On the basis of the clinical experience of Ortho-CT, a new style of limited cone-beam X-CT called ''3DX'' was developed by us for practical use. The skin dose is almost the same as in the prototype system. The images show very high resolution compared to those of the prototype system. We think that this system is very useful for diagnosis of hard tissue for

  13. Utility of cone-beam CT imaging in prostatic artery embolization.

    Science.gov (United States)

    Bagla, Sandeep; Rholl, Kenneth S; Sterling, Keith M; van Breda, Arletta; Papadouris, Dimitrios; Cooper, James M; van Breda, Arina

    2013-11-01

    To evaluate the utility of cone-beam computed tomography (CT) in patients undergoing prostatic artery (PA) embolization (PAE) for benign prostatic hyperplasia. From January 2012 to January 2013, 15 patients (age range, 59-81 y; mean, 68 y) with moderate- or severe-grade lower urinary tract symptoms, in whom medical management had failed were enrolled in a prospective United States trial to evaluate PAE. During pelvic angiography, 15 cone-beam CT acquisitions were performed in 11 patients, and digital subtraction angiography was performed in all patients. Cone-beam CT images were reviewed to assess for sites of potential nontarget embolization that impacted therapy, a pattern of enhancement on cone-beam CT suggesting additional PAs, confirmation of prostatic parenchymal perfusion before embolization, and contralateral prostatic parenchymal enhancement. Cone-beam CT was successful in 14 of 15 acquisitions, and PAE was successful in 14 of 15 patients (92%). Cone-beam CT provided information that impacted treatment in five of 11 patients (46%) by allowing for identification of sites of potential nontarget embolization. Duplicated prostatic arterial supply and contralateral perfusion were each identified in 21% of patients (three of 11). Prostatic perfusion was confirmed before embolization in 50% of acquisitions (seven of 14). Cone-beam CT is a useful technique that can potentially mitigate the risk of nontarget embolization. During treatment, it can allow for the interventionalist to identify duplicated prostatic arterial supply or contralateral perfusion, which may be useful when evaluating a treatment failure. © SIR, 2013.

  14. Reducing radiation dose to the female breast during CT coronary angiography: A simulation study comparing breast shielding, angular tube current modulation, reduced kV, and partial angle protocols using an unknown-location signal-detectability metric

    Energy Technology Data Exchange (ETDEWEB)

    Rupcich, Franco; Gilat Schmidt, Taly [Department of Biomedical Engineering, Marquette University, Milwaukee, Wisconsin 53233 (United States); Badal, Andreu; Popescu, Lucretiu M.; Kyprianou, Iacovos [Center for Devices and Radiological Health, US Food and Drug Administration, Silver Spring, Maryland 20905 (United States)

    2013-08-15

    Purpose: The authors compared the performance of five protocols intended to reduce dose to the breast during computed tomography (CT) coronary angiography scans using a model observer unknown-location signal-detectability metric.Methods: The authors simulated CT images of an anthropomorphic female thorax phantom for a 120 kV reference protocol and five “dose reduction” protocols intended to reduce dose to the breast: 120 kV partial angle (posteriorly centered), 120 kV tube-current modulated (TCM), 120 kV with shielded breasts, 80 kV, and 80 kV partial angle (posteriorly centered). Two image quality tasks were investigated: the detection and localization of 4-mm, 3.25 mg/ml and 1-mm, 6.0 mg/ml iodine contrast signals randomly located in the heart region. For each protocol, the authors plotted the signal detectability, as quantified by the area under the exponentially transformed free response characteristic curve estimator (A-caret{sub FE}), as well as noise and contrast-to-noise ratio (CNR) versus breast and lung dose. In addition, the authors quantified each protocol's dose performance as the percent difference in dose relative to the reference protocol achieved while maintaining equivalent A-caret{sub FE}.Results: For the 4-mm signal-size task, the 80 kV full scan and 80 kV partial angle protocols decreased dose to the breast (80.5% and 85.3%, respectively) and lung (80.5% and 76.7%, respectively) with A-caret{sub FE} = 0.96, but also resulted in an approximate three-fold increase in image noise. The 120 kV partial protocol reduced dose to the breast (17.6%) at the expense of increased lung dose (25.3%). The TCM algorithm decreased dose to the breast (6.0%) and lung (10.4%). Breast shielding increased breast dose (67.8%) and lung dose (103.4%). The 80 kV and 80 kV partial protocols demonstrated greater dose reductions for the 4-mm task than for the 1-mm task, and the shielded protocol showed a larger increase in dose for the 4-mm task than for the 1-mm

  15. Determination of dose enhancement caused by gold-nanoparticles irradiated with proton, X-rays (kV and MV) and electron beams, using alanine/EPR dosimeters

    International Nuclear Information System (INIS)

    Smith, Clare L.; Ackerly, Trevor; Best, Stephen P.; Gagliardi, Frank; Kie, Katahira; Little, Peter J.; McCorkell, Giulia; Sale, Charlotte A.; Tsunei, Yusuke; Tominaga, Takahiro; Volaric, Sioe See; Geso, Moshi

    2015-01-01

    The main aims of this research was to employ alanine doped with gold-nanoparticles “AuNPs” to determine the levels of dose enhancement caused by these particles when irradiated with proton beams, low and high energy X-rays and electrons. DL-alanine was impregnated with 5 nm gold-nanoparticles (3% by weight) and added as a uniform layer within a wax pellet of dimensions 10 × 5 × 5 mm. Control pellets, containing DL-Alanine were also produced, and placed within a phantom, and exposed to various types of radiations: low energy (kV ranges) X-rays were obtained from a superficial machine, high energy (MV) X-rays and electrons derived from a linear accelerator, and protons were produced by the Hyogo Ion Beam Centre in Japan. Nominal doses received ranged from 2 to 20 Gy (within clinical range). The Electron Paramagnetic Resonance (EPR) spectra of the irradiated samples were recorded on a BRUKER Elexsys 9.5 MHz. The dose enhancement caused by gold nanoparticles for 80 kV x-rays was found to be more than 60% at about 5 Gy. Smaller dose enhancements (under the same measurement conditions) were observed for megavoltage x-ray beams (up to 10%). Dose enhancement caused by charged particles indicated minimal values for 6 MeV electrons (approximately 5%) whilst less than that is obtained with protons of 150 MeV. The proton results validate the latest simulation results based on Monte Carlo calculations but the dose enhancement is significantly less than that reported in cell and animal model systems, (about 20%). We attribute this difference to the fact that alanine only measures the levels of free radicals generated by the inclusion of nanoparticles and not the redox type radicals (such as reactive oxygen species) generated from aqueous media in cells. Dose enhancement caused by 5 nm gold-nanoparticles with radiotherapy type proton beams has been found to be less than 5% as determined when using alanine/wax as both a phantom and dosimeter. This agrees well

  16. Noise simulation in cone beam CT imaging with parallel computing

    International Nuclear Information System (INIS)

    Tu, S.-J.; Shaw, Chris C; Chen, Lingyun

    2006-01-01

    We developed a computer noise simulation model for cone beam computed tomography imaging using a general purpose PC cluster. This model uses a mono-energetic x-ray approximation and allows us to investigate three primary performance components, specifically quantum noise, detector blurring and additive system noise. A parallel random number generator based on the Weyl sequence was implemented in the noise simulation and a visualization technique was accordingly developed to validate the quality of the parallel random number generator. In our computer simulation model, three-dimensional (3D) phantoms were mathematically modelled and used to create 450 analytical projections, which were then sampled into digital image data. Quantum noise was simulated and added to the analytical projection image data, which were then filtered to incorporate flat panel detector blurring. Additive system noise was generated and added to form the final projection images. The Feldkamp algorithm was implemented and used to reconstruct the 3D images of the phantoms. A 24 dual-Xeon PC cluster was used to compute the projections and reconstructed images in parallel with each CPU processing 10 projection views for a total of 450 views. Based on this computer simulation system, simulated cone beam CT images were generated for various phantoms and technique settings. Noise power spectra for the flat panel x-ray detector and reconstructed images were then computed to characterize the noise properties. As an example among the potential applications of our noise simulation model, we showed that images of low contrast objects can be produced and used for image quality evaluation

  17. Surgical stent for dental implant using cone beam CT images

    International Nuclear Information System (INIS)

    Choi, Hyung Soo; Kim, Gyu Tae; Choi, Yong Suk; Hwang, Eui Hwan

    2010-01-01

    The purpose of this study is to develop a surgical stent for dental implant procedure that can be easily applied and affordable by using cone beam computerized tomography (CBCT). Aluminum, Teflon-PFA (perfluoroalkoxy), and acetal (polyoxymethylene plastic) were selected as materials for the surgical stent. Among these three materials, the appropriate material was chosen using the CBCT images. The surgical stent, which could be easily placed into an oral cavity, was designed with chosen material. CBCT images of the new surgical stent on mandible were obtained using Alphard-3030 dental CT system (Asahi Roentgen Co., Ltd., Kyoto, Japan). The point of insertion was prescribed on the surgical stent with the multiplanar reconstruction software of OnDemand3D (CyberMed Inc., Seoul, Korea). Guide holes were made at the point of insertion on the surgical stent using newly designed guide jig. CBCT scans was taken for the second time to verify the accuracy of the newly designed surgical stent. Teflon-PFA showed radiologically excellent image characteristics for the surgical stent. High accuracy and reproducibility of implantation were confirmed with the surgical stent. The newly designed surgical stent can lead to the accurate implantation and achieve the clinically predictable result.

  18. Surgical stent for dental implant using cone beam CT images

    Energy Technology Data Exchange (ETDEWEB)

    Choi, Hyung Soo; Kim, Gyu Tae; Choi, Yong Suk; Hwang, Eui Hwan [Department of Oral and Maxillofacial Radiology, School of Dentistry, Kung Hee University, Seoul (Korea, Republic of)

    2010-12-15

    The purpose of this study is to develop a surgical stent for dental implant procedure that can be easily applied and affordable by using cone beam computerized tomography (CBCT). Aluminum, Teflon-PFA (perfluoroalkoxy), and acetal (polyoxymethylene plastic) were selected as materials for the surgical stent. Among these three materials, the appropriate material was chosen using the CBCT images. The surgical stent, which could be easily placed into an oral cavity, was designed with chosen material. CBCT images of the new surgical stent on mandible were obtained using Alphard-3030 dental CT system (Asahi Roentgen Co., Ltd., Kyoto, Japan). The point of insertion was prescribed on the surgical stent with the multiplanar reconstruction software of OnDemand3D (CyberMed Inc., Seoul, Korea). Guide holes were made at the point of insertion on the surgical stent using newly designed guide jig. CBCT scans was taken for the second time to verify the accuracy of the newly designed surgical stent. Teflon-PFA showed radiologically excellent image characteristics for the surgical stent. High accuracy and reproducibility of implantation were confirmed with the surgical stent. The newly designed surgical stent can lead to the accurate implantation and achieve the clinically predictable result.

  19. Comparison of Image Quality and Diagnostic Performance of Cone-Beam CT during Drug-Eluting Embolic Transarterial Chemoembolization and Multidetector CT in the Detection of Hepatocellular Carcinoma.

    Science.gov (United States)

    Lucatelli, Pierleone; Argirò, Renato; Ginanni Corradini, Stefano; Saba, Luca; Cirelli, Carlo; Fanelli, Fabrizio; Ricci, Carmelo; Levi Sandri, Giovanni Battista; Catalano, Carlo; Bezzi, Mario

    2017-07-01

    To compare image quality and diagnostic performance of cone-beam computed tomography (CT) and multidetector CT in the detection of hypervascular hepatocellular carcinoma (HCC) in patients with cirrhosis undergoing transarterial chemoembolization with drug-eluting embolic agents. Fifty-five consecutive patients referred for chemoembolization of hypervascular HCC were prospectively enrolled. Imaging included preprocedural multidetector CT within 1 month before planned treatment, intraprocedural cone-beam CT, and 1-month follow-up multidetector CT. Analysis of image quality was performed with calculations of lesion-to-liver contrast-to-noise ratio (LLCNR) and lesion-to-liver signal-to-noise-ratio (LLSNR). One-month follow-up multidetector CT was considered the reference standard for the detection of HCC nodules. Median LLCNR values were 3.94 (95% confidence interval [CI], 3.06-5.05) for preprocedural multidetector CT and 6.90 (95% CI, 5.17-7.77) for intraprocedural cone-beam CT (P cone-beam CT (P cone-beam CT detected 15 additional hypervascular nodules that were also visible on 1-month follow-up multidetector CT. Cone-beam CT has a significantly higher diagnostic performance compared with preprocedural multidetector CT in the detection of HCCs and can influence management of patients with cirrhosis by identifying particularly aggressive tumors. Copyright © 2017 SIR. Published by Elsevier Inc. All rights reserved.

  20. SU-F-J-114: On-Treatment Imagereconstruction Using Transit Images of Treatment Beams Through Patient and Thosethrough Planning CT Images

    Energy Technology Data Exchange (ETDEWEB)

    Lee, H; Cho, S [KAIST, Yuseong-gu, Daejeon (Korea, Republic of); Cheong, K [Hallym University Sacred Heart Hospital, Anyang (Korea, Republic of); Jung, J [East Carolina University Greenville, NC (United States); Jung, S [Samsung Medical Cener, Gangnam-gu, Seoul (Korea, Republic of); Kim, J [Yonsei Cancer Center, Seoul (Korea, Republic of); Yeo, I [Loma Linda University Medical Center, Loma Linda, CA (United States)

    2016-06-15

    Purpose: To reconstruct patient images at the time of radiation delivery using measured transit images of treatment beams through patient and calculated transit images through planning CT images. Methods: We hypothesize that the ratio of the measured transit images to the calculated images may provide changed amounts of the patient image between times of planning CT and treatment. To test, we have devised lung phantoms with a tumor object (3-cm diameter) placed at iso-center (simulating planning CT) and off-center by 1 cm (simulating treatment). CT images of the two phantoms were acquired; the image of the off-centered phantom, unavailable clinically, represents the reference on-treatment image in the image quality of planning CT. Cine-transit images through the two phantoms were also acquired in EPID from a non-modulated 6 MV beam when the gantry was rotated 360 degrees; the image through the centered phantom simulates calculated image. While the current study is a feasibility study, in reality our computational EPID model can be applicable in providing accurate transit image from MC simulation. Changed MV HU values were reconstructed from the ratio between two EPID projection data, converted to KV HU values, and added to the planning CT, thereby reconstructing the on-treatment image of the patient limited to the irradiated region of the phantom. Results: The reconstructed image was compared with the reference image. Except for local HU differences>200 as a maximum, excellent agreement was found. The average difference across the entire image was 16.2 HU. Conclusion: We have demonstrated the feasibility of a method of reconstructing on-treatment images of a patient using EPID image and planning CT images. Further studies will include resolving the local HU differences and investigation on the dosimetry impact of the reconstructed image.

  1. State-of-the-art on cone beam CT imaging for preoperative planning of implant placement.

    NARCIS (Netherlands)

    Guerrero, M.E.; Jacobs, R.; Loubele, M.; Schutyser, F.A.C.; Suetens, P.; Steenberghe, D van

    2006-01-01

    Orofacial diagnostic imaging has grown dramatically in recent years. As the use of endosseous implants has revolutionized oral rehabilitation, a specialized technique has become available for the preoperative planning of oral implant placement: cone beam computed tomography (CT). This imaging

  2. Determination of size-specific exposure settings in dental cone-beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Pauwels, Ruben [Chulalongkorn University, Department of Radiology, Faculty of Dentistry, Patumwan, Bangkok (Thailand); University of Leuven, OMFS-IMPATH Research Group, Department of Imaging and Pathology, Biomedical Sciences Group, Leuven (Belgium); Jacobs, Reinhilde [University of Leuven, OMFS-IMPATH Research Group, Department of Imaging and Pathology, Biomedical Sciences Group, Leuven (Belgium); Bogaerts, Ria [University of Leuven, Laboratory of Experimental Radiotherapy, Department of Oncology, Biomedical Sciences Group, Leuven (Belgium); Bosmans, Hilde [University of Leuven, Medical Physics and Quality Assessment, Department of Imaging and Pathology, Biomedical Sciences Group, Leuven (Belgium); Panmekiate, Soontra [Chulalongkorn University, Department of Radiology, Faculty of Dentistry, Patumwan, Bangkok (Thailand)

    2017-01-15

    To estimate the possible reduction of tube output as a function of head size in dental cone-beam computed tomography (CBCT). A 16 cm PMMA phantom, containing a central and six peripheral columns filled with PMMA, was used to represent an average adult male head. The phantom was scanned using CBCT, with 0-6 peripheral columns having been removed in order to simulate varying head sizes. For five kV settings (70-90 kV), the mAs required to reach a predetermined image noise level was determined, and corresponding radiation doses were derived. Results were expressed as a function of head size, age, and gender, based on growth reference charts. The use of 90 kV consistently resulted in the largest relative dose reduction. A potential mAs reduction ranging from 7 % to 50 % was seen for the different simulated head sizes, showing an exponential relation between head size and mAs. An optimized exposure protocol based on head circumference or age/gender is proposed. A considerable dose reduction, through reduction of the mAs rather than the kV, is possible for small-sized patients in CBCT, including children and females. Size-specific exposure protocols should be clinically implemented. (orig.)

  3. Determination of size-specific exposure settings in dental cone-beam CT

    International Nuclear Information System (INIS)

    Pauwels, Ruben; Jacobs, Reinhilde; Bogaerts, Ria; Bosmans, Hilde; Panmekiate, Soontra

    2017-01-01

    To estimate the possible reduction of tube output as a function of head size in dental cone-beam computed tomography (CBCT). A 16 cm PMMA phantom, containing a central and six peripheral columns filled with PMMA, was used to represent an average adult male head. The phantom was scanned using CBCT, with 0-6 peripheral columns having been removed in order to simulate varying head sizes. For five kV settings (70-90 kV), the mAs required to reach a predetermined image noise level was determined, and corresponding radiation doses were derived. Results were expressed as a function of head size, age, and gender, based on growth reference charts. The use of 90 kV consistently resulted in the largest relative dose reduction. A potential mAs reduction ranging from 7 % to 50 % was seen for the different simulated head sizes, showing an exponential relation between head size and mAs. An optimized exposure protocol based on head circumference or age/gender is proposed. A considerable dose reduction, through reduction of the mAs rather than the kV, is possible for small-sized patients in CBCT, including children and females. Size-specific exposure protocols should be clinically implemented. (orig.)

  4. [Development of a cone-beam CT system for radiological technologist education].

    Science.gov (United States)

    Teramoto, Atsushi; Ozaki, Kaho; Miyashita, Mariko; Ohno, Tomoyuki; Tsuzaka, Masatoshi; Fujita, Hiroshi; Ohara, Ken

    2011-01-01

    For radiological technologists, it is very important to understand the principle of computed tomography (CT) and CT artifacts derived from mechanical and electrical failure. In this study, a CT system for educating radiological technologists was developed. The system consisted of a cone-beam CT scanner and educational software. The cone-beam CT scanner has a simple structure, using a micro-focus X-ray tube and an indirect-conversion flat panel detector. For the educational software, we developed various educational functions of image reconstruction and reconstruction parameters as well as CT artifacts. In the experiments, the capabilities of the system were evaluated using an acrylic phantom. We verified that the system produced the expected results.

  5. Development of a cone-beam CT system for radiological technologist education

    International Nuclear Information System (INIS)

    Teramoto, Atsushi; Ohara, Ken; Ozaki, Kaho; Miyashita, Mariko; Ohno, Tomoyuki; Tsuzaka, Masatoshi; Fujita, Hiroshi

    2011-01-01

    For radiological technologists, it is very important to understand the principle of computed tomography (CT) and CT artifacts derived from mechanical and electrical failure. In this study, a CT system for educating radiological technologists was developed. The system consisted of a cone-beam CT scanner and educational software. The cone-beam CT scanner has a simple structure, using a micro-focus X-ray tube and an indirect-conversion flat panel detector. For the educational software, we developed various educational functions of image reconstruction and reconstruction parameters as well as CT artifacts. In the experiments, the capabilities of the system were evaluated using an acrylic phantom. We verified that the system produced the expected results. (author)

  6. Noise power properties of a cone-beam CT system for breast cancer detection

    OpenAIRE

    Yang, Kai; Kwan, Alexander L.C.; Huang, Shih-Ying; Packard, Nathan J.; Boone, John M.

    2008-01-01

    The noise power properties of a cone-beam computed tomography (CT) system dedicated for breast cancer detection were investigated. Uniform polyethylene cylinders of various diameters were scanned under different system acquisition conditions. Noise power spectra were calculated from difference data generated by subtraction between two identical scans. Multidimensional noise power spectra (NPS) were used as the metric to evaluate the noise properties of the breast CT (bCT) under different syst...

  7. 4D cone beam CT via spatiotemporal tensor framelet

    International Nuclear Information System (INIS)

    Gao, Hao; Li, Ruijiang; Xing, Lei; Lin, Yuting

    2012-01-01

    Purpose: On-board 4D cone beam CT (4DCBCT) offers respiratory phase-resolved volumetric imaging, and improves the accuracy of target localization in image guided radiation therapy. However, the clinical utility of this technique has been greatly impeded by its degraded image quality, prolonged imaging time, and increased imaging dose. The purpose of this letter is to develop a novel iterative 4DCBCT reconstruction method for improved image quality, increased imaging speed, and reduced imaging dose. Methods: The essence of this work is to introduce the spatiotemporal tensor framelet (STF), a high-dimensional tensor generalization of the 1D framelet for 4DCBCT, to effectively take into account of highly correlated and redundant features of the patient anatomy during respiration, in a multilevel fashion with multibasis sparsifying transform. The STF-based algorithm is implemented on a GPU platform for improved computational efficiency. To evaluate the method, 4DCBCT full-fan scans were acquired within 30 s, with a gantry rotation of 200°; STF is also compared with a state-of-art reconstruction method via spatiotemporal total variation regularization. Results: Both the simulation and experimental results demonstrate that STF-based reconstruction achieved superior image quality. The reconstruction of 20 respiratory phases took less than 10 min on an NVIDIA Tesla C2070 GPU card. The STF codes are available at https://sites.google.com/site/spatiotemporaltensorframelet . Conclusions: By effectively utilizing the spatiotemporal coherence of the patient anatomy among different respiratory phases in a multilevel fashion with multibasis sparsifying transform, the proposed STF method potentially enables fast and low-dose 4DCBCT with improved image quality.

  8. Study of residual stresses in CT test specimens welded by electron beam

    Science.gov (United States)

    Papushkin, I. V.; Kaisheva, D.; Bokuchava, G. D.; Angelov, V.; Petrov, P.

    2018-03-01

    The paper reports result of residual stress distribution studies in CT specimens reconstituted by electron beam welding (EBW). The main aim of the study is evaluation of the applicability of the welding technique for CT specimens’ reconstitution. Thus, the temperature distribution during electron beam welding of a CT specimen was calculated using Green’s functions and the residual stress distribution was determined experimentally using neutron diffraction. Time-of-flight neutron diffraction experiments were performed on a Fourier stress diffractometer at the IBR-2 fast pulsed reactor in FLNP JINR (Dubna, Russia). The neutron diffraction data estimates yielded a maximal stress level of ±180 MPa in the welded joint.

  9. Evaluation of positioning errors of the patient using cone beam CT megavoltage; Evaluacion de errores de posicionamiento del paciente mediante Cone Beam CT de megavoltaje

    Energy Technology Data Exchange (ETDEWEB)

    Garcia Ruiz-Zorrilla, J.; Fernandez Leton, J. P.; Zucca Aparicio, D.; Perez Moreno, J. M.; Minambres Moro, A.

    2013-07-01

    Image-guided radiation therapy allows you to assess and fix the positioning of the patient in the treatment unit, thus reducing the uncertainties due to the positioning of the patient. This work assesses errors systematic and errors of randomness from the corrections made to a series of patients of different diseases through a protocol off line of cone beam CT (CBCT) megavoltage. (Author)

  10. Baseball II-T multiple beam project 50 kV, 80 amp, power supply system for sustaining neutral beam sources

    International Nuclear Information System (INIS)

    Waugh, A.F.

    1976-01-01

    This document describes a high power, high voltage power supply system required to operate sustaining neutral beam (NB) sources as would be used in the Baseball II-T Multiple Beam Project. Two such systems would be needed. Appended are site plan installation drawings

  11. Diagnosis of pulmonary embolism with spiral and electron-beam CT; Diagnostik der Lungenembolie mit Spiral- und Elektronenstrahl-CT

    Energy Technology Data Exchange (ETDEWEB)

    Schoepf, U.J.; Bruening, R.D.; Becker, C.R.; Konschitzky, H.; Muehling, O.; Staebler, A.; Helmberger, T.; Holzknecht, N.; Reiser, M.F. [Muenchen Univ. (Germany). Inst. fuer Radiologische Diagnostik; Knez, A.; Haberl, R. [Muenchen Univ. (Germany). Medizinische Klinik 1

    1998-12-01

    Purpose: To compare spiral (SCT) and electron-beam CT (EBT) for the diagnosis of pulmonary embolism (PE). Materials and methods: From June 1997 to June 1998 188 patients with suspected acute or chronic thrombembolism of the pulmonary arteries were examined. A total of 108 patients were scanned using SCT and 80 patients using EBT. On each scanner two different scan protocols were evaluated. Conclusions: Advanced CT scanning techniques allow the highly accurate diagnosis of central and peripheral PE. Other potentially life-threatening underlying diseases are also readily recognized. (orig./AJ) [Deutsch] Fragestellung: Spiral-CT (SCT) und Elektronenstrahlcomputertomographie (EBT) sollten hinsichtlich ihrer Eignung fuer die Diagnostik der Lungenembolie (LE) verglichen werden. Methode: Von Juni 1997 bis Juni 1998 wurden 188 Patienten mit Verdacht auf akute oder chronische thrombembolische Veraenderungen der Lungenarterien untersucht. Die CT-Diagnostik erfolgte dabei bei 108 Patienten mit Spiral-CT und bei 80 Patienten mit EBT. Schlussfolgerungen: Moderne CT-Scan-Verfahren erlauben mit hoher Genauigkeit die Diagnose der zentralen und peripheren Lungenembolie. Die EBT bietet Vorteile in der Darstellung herznaher peripherer Lungenarterien. Andere lebendsbedrohliche Ursachen fuer die Beschwerden des Patienten werden mit der CT sicher erkannt. (orig./AJ)

  12. Identification of dental root canals and their medial line from micro-CT and cone-beam CT records

    Directory of Open Access Journals (Sweden)

    Benyó Balázs

    2012-10-01

    Full Text Available Abstract Background Shape of the dental root canal is highly patient specific. Automated identification methods of the medial line of dental root canals and the reproduction of their 3D shape can be beneficial for planning endodontic interventions as severely curved root canals or multi-rooted teeth may pose treatment challenges. Accurate shape information of the root canals may also be used by manufacturers of endodontic instruments in order to make more efficient clinical tools. Method Novel image processing procedures dedicated to the automated detection of the medial axis of the root canal from dental micro-CT and cone-beam CT records are developed. For micro-CT, the 3D model of the root canal is built up from several hundred parallel cross sections, using image enhancement, histogram based fuzzy c-means clustering, center point detection in the segmented slice, three dimensional inner surface reconstruction, and potential field driven curve skeleton extraction in three dimensions. Cone-beam CT records are processed with image enhancement filters and fuzzy chain based regional segmentation, followed by the reconstruction of the root canal surface and detecting its skeleton via a mesh contraction algorithm. Results The proposed medial line identification and root canal detection algorithms are validated on clinical data sets. 25 micro-CT and 36 cone-beam-CT records are used in the validation procedure. The overall success rate of the automatic dental root canal identification was about 92% in both procedures. The algorithms proved to be accurate enough for endodontic therapy planning. Conclusions Accurate medial line identification and shape detection algorithms of dental root canal have been developed. Different procedures are defined for micro-CT and cone-beam CT records. The automated execution of the subsequent processing steps allows easy application of the algorithms in the dental care. The output data of the image processing procedures

  13. Benign Prostatic Hyperplasia: Cone-Beam CT in Conjunction with DSA for Identifying Prostatic Arterial Anatomy.

    Science.gov (United States)

    Wang, Mao Qiang; Duan, Feng; Yuan, Kai; Zhang, Guo Dong; Yan, Jieyu; Wang, Yan

    2017-01-01

    Purpose To describe findings in prostatic arteries (PAs) at digital subtraction angiography (DSA) and cone-beam computed tomography (CT) that allow identification of benign prostatic hyperplasia and to determine the value added with the use of cone-beam CT. Materials and Methods This retrospective single-institution study was approved by the institutional review board, and the requirement for written informed consent was waived. From February 2009 to December 2014, a total of 148 patients (mean age ± standard deviation, 70.5 years ± 14.5) underwent DSA of the internal iliac arteries and cone-beam CT with a flat-detector angiographic system before they underwent prostate artery embolization. Both the DSA and cone-beam CT images were evaluated by two interventional radiologists to determine the number of independent PAs and their origins and anastomoses with adjacent arteries. The exact McNemar test was used to compare the detection rate of the PAs and the anastomoses with DSA and with cone-beam CT. Results The PA anatomy was evaluated successfully by means of cone-beam CT in conjunction with DSA in all patients. Of the 296 pelvic sides, 274 (92.6%) had only one PA. The most frequent PA origin was the common gluteal-pudendal trunk with the superior vesicular artery in 118 (37.1%), followed by the anterior division of the internal iliac artery in 99 (31.1%), and the internal pudendal artery in 77 (24.2%) pelvic sides. In 67 (22.6%) pelvic sides, anastomoses to adjacent arteries were documented. The numbers of PA origins and anastomoses, respectively, that could be identified were significantly higher with cone-beam CT (301 of 318 [94.7%] and 65 of 67 [97.0%]) than with DSA (237 [74.5%] and 39 [58.2%], P Cone-beam CT provided essential information that was not available with DSA in 90 of 148 (60.8%) patients. Conclusion Cone-beam CT is a useful adjunctive technique to DSA for identification of the PA anatomy and provides information to help treatment planning during

  14. Evaluation of the OSC-TV iterative reconstruction algorithm for cone-beam optical CT

    Energy Technology Data Exchange (ETDEWEB)

    Matenine, Dmitri, E-mail: dmitri.matenine.1@ulaval.ca; Mascolo-Fortin, Julia, E-mail: julia.mascolo-fortin.1@ulaval.ca [Département de physique, de génie physique et d’optique, Université Laval, Québec, Québec G1V 0A6 (Canada); Goussard, Yves, E-mail: yves.goussard@polymtl.ca [Département de génie électrique/Institut de génie biomédical, École Polytechnique de Montréal, C.P. 6079, succ. Centre-ville, Montréal, Québec H3C 3A7 (Canada); Després, Philippe, E-mail: philippe.despres@phy.ulaval.ca [Département de physique, de génie physique et d’optique and Centre de recherche sur le cancer, Université Laval, Québec, Québec G1V 0A6, Canada and Département de radio-oncologie and Centre de recherche du CHU de Québec, Québec, Québec G1R 2J6 (Canada)

    2015-11-15

    Purpose: The present work evaluates an iterative reconstruction approach, namely, the ordered subsets convex (OSC) algorithm with regularization via total variation (TV) minimization in the field of cone-beam optical computed tomography (optical CT). One of the uses of optical CT is gel-based 3D dosimetry for radiation therapy, where it is employed to map dose distributions in radiosensitive gels. Model-based iterative reconstruction may improve optical CT image quality and contribute to a wider use of optical CT in clinical gel dosimetry. Methods: This algorithm was evaluated using experimental data acquired by a cone-beam optical CT system, as well as complementary numerical simulations. A fast GPU implementation of OSC-TV was used to achieve reconstruction times comparable to those of conventional filtered backprojection. Images obtained via OSC-TV were compared with the corresponding filtered backprojections. Spatial resolution and uniformity phantoms were scanned and respective reconstructions were subject to evaluation of the modulation transfer function, image uniformity, and accuracy. The artifacts due to refraction and total signal loss from opaque objects were also studied. Results: The cone-beam optical CT data reconstructions showed that OSC-TV outperforms filtered backprojection in terms of image quality, thanks to a model-based simulation of the photon attenuation process. It was shown to significantly improve the image spatial resolution and reduce image noise. The accuracy of the estimation of linear attenuation coefficients remained similar to that obtained via filtered backprojection. Certain image artifacts due to opaque objects were reduced. Nevertheless, the common artifact due to the gel container walls could not be eliminated. Conclusions: The use of iterative reconstruction improves cone-beam optical CT image quality in many ways. The comparisons between OSC-TV and filtered backprojection presented in this paper demonstrate that OSC-TV can

  15. Evaluation of the OSC-TV iterative reconstruction algorithm for cone-beam optical CT.

    Science.gov (United States)

    Matenine, Dmitri; Mascolo-Fortin, Julia; Goussard, Yves; Després, Philippe

    2015-11-01

    The present work evaluates an iterative reconstruction approach, namely, the ordered subsets convex (OSC) algorithm with regularization via total variation (TV) minimization in the field of cone-beam optical computed tomography (optical CT). One of the uses of optical CT is gel-based 3D dosimetry for radiation therapy, where it is employed to map dose distributions in radiosensitive gels. Model-based iterative reconstruction may improve optical CT image quality and contribute to a wider use of optical CT in clinical gel dosimetry. This algorithm was evaluated using experimental data acquired by a cone-beam optical CT system, as well as complementary numerical simulations. A fast GPU implementation of OSC-TV was used to achieve reconstruction times comparable to those of conventional filtered backprojection. Images obtained via OSC-TV were compared with the corresponding filtered backprojections. Spatial resolution and uniformity phantoms were scanned and respective reconstructions were subject to evaluation of the modulation transfer function, image uniformity, and accuracy. The artifacts due to refraction and total signal loss from opaque objects were also studied. The cone-beam optical CT data reconstructions showed that OSC-TV outperforms filtered backprojection in terms of image quality, thanks to a model-based simulation of the photon attenuation process. It was shown to significantly improve the image spatial resolution and reduce image noise. The accuracy of the estimation of linear attenuation coefficients remained similar to that obtained via filtered backprojection. Certain image artifacts due to opaque objects were reduced. Nevertheless, the common artifact due to the gel container walls could not be eliminated. The use of iterative reconstruction improves cone-beam optical CT image quality in many ways. The comparisons between OSC-TV and filtered backprojection presented in this paper demonstrate that OSC-TV can potentially improve the rendering of

  16. Evaluation of the OSC-TV iterative reconstruction algorithm for cone-beam optical CT

    International Nuclear Information System (INIS)

    Matenine, Dmitri; Mascolo-Fortin, Julia; Goussard, Yves; Després, Philippe

    2015-01-01

    Purpose: The present work evaluates an iterative reconstruction approach, namely, the ordered subsets convex (OSC) algorithm with regularization via total variation (TV) minimization in the field of cone-beam optical computed tomography (optical CT). One of the uses of optical CT is gel-based 3D dosimetry for radiation therapy, where it is employed to map dose distributions in radiosensitive gels. Model-based iterative reconstruction may improve optical CT image quality and contribute to a wider use of optical CT in clinical gel dosimetry. Methods: This algorithm was evaluated using experimental data acquired by a cone-beam optical CT system, as well as complementary numerical simulations. A fast GPU implementation of OSC-TV was used to achieve reconstruction times comparable to those of conventional filtered backprojection. Images obtained via OSC-TV were compared with the corresponding filtered backprojections. Spatial resolution and uniformity phantoms were scanned and respective reconstructions were subject to evaluation of the modulation transfer function, image uniformity, and accuracy. The artifacts due to refraction and total signal loss from opaque objects were also studied. Results: The cone-beam optical CT data reconstructions showed that OSC-TV outperforms filtered backprojection in terms of image quality, thanks to a model-based simulation of the photon attenuation process. It was shown to significantly improve the image spatial resolution and reduce image noise. The accuracy of the estimation of linear attenuation coefficients remained similar to that obtained via filtered backprojection. Certain image artifacts due to opaque objects were reduced. Nevertheless, the common artifact due to the gel container walls could not be eliminated. Conclusions: The use of iterative reconstruction improves cone-beam optical CT image quality in many ways. The comparisons between OSC-TV and filtered backprojection presented in this paper demonstrate that OSC-TV can

  17. CT calibration for two-dimensional scaling of proton pencil beams

    International Nuclear Information System (INIS)

    Szymanowski, Hanitra; Oelfke, Uwe

    2003-01-01

    For proton dose calculations in heterogeneous media, it was shown in a previous work that the conventional pencil beam approach based on pathlength scaling does not properly account for scattering effects in nonwater media (Szymanowski and Oelfke 2002 Phys. Med. Biol. 47 3313-30). A two-dimensional scaling method was therefore introduced, which is able to predict with high accuracy the propagation of proton pencil beams both along the depth and the lateral directions in inhomogeneous media. In order to integrate this improved pencil beam algorithm in a CT based treatment planning system, two CT calibration curves are needed. The first one relates the Hounsfield numbers to the relative stopping powers, as for the conventional pencil beam approach. The second curve is to relate the Hounsfield numbers to the material-specific lateral scaling factors. The purpose of this work is to provide the CT calibration curves needed for the integration of the pencil beam algorithm featuring the two-dimensional scaling method. Similarly to as suggested by Schneider et al (1996 Phys. Med. Biol. 41 111-24) for the calibration curve in terms of stopping powers, we follow a stoichiometric procedure to get the calibration curve in terms of material-specific lateral scaling factors. The calibration curves for a CT scanner of the type Siemens Somatom Plus 4 are obtained from the analytical calculation of the CT Hounsfield numbers, relative stopping powers and material-specific lateral scaling factors for human biological tissues

  18. Metal Artifact Reduction for Polychromatic X-ray CT Based on a Beam-Hardening Corrector.

    Science.gov (United States)

    Park, Hyoung Suk; Hwang, Dosik; Seo, Jin Keun

    2016-02-01

    This paper proposes a new method to correct beam hardening artifacts caused by the presence of metal in polychromatic X-ray computed tomography (CT) without degrading the intact anatomical images. Metal artifacts due to beam-hardening, which are a consequence of X-ray beam polychromaticity, are becoming an increasingly important issue affecting CT scanning as medical implants become more common in a generally aging population. The associated higher-order beam-hardening factors can be corrected via analysis of the mismatch between measured sinogram data and the ideal forward projectors in CT reconstruction by considering the known geometry of high-attenuation objects. Without prior knowledge of the spectrum parameters or energy-dependent attenuation coefficients, the proposed correction allows the background CT image (i.e., the image before its corruption by metal artifacts) to be extracted from the uncorrected CT image. Computer simulations and phantom experiments demonstrate the effectiveness of the proposed method to alleviate beam hardening artifacts.

  19. Three dimensional evaluation of impacted mesiodens using dental cone beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Dong Ho; Lee, Jae Seo; Yoon, Suk Ja; Kang, Byung Cheol [Chonnam National University School of Medicine, Gwangju (Korea, Republic of)

    2010-09-15

    This study was performed to analyze the position, pattern of impacted mesiodens, and their relationship to the adjacent teeth using Dental cone-beam CT. Sixty-two dental cone-beam CT images with 81 impacted mesiodenses were selected from about 2,298 cone-beam CT images at Chonnam National University Dental Hospital from June 2006 to March 2009. The position, pattern, shape of impacted mesiodenses and their complications were analyzed in cone-beam CT including 3D images. The sex ratio (M : F) was 2.9 : 1. Most of the mesiodenses (87.7%) were located at palatal side to the incisors. 79% of the mesiodenses were conical in shape. 60.5% of the mesiodenses were inverted, 21% normal erupting direction, and 18.5% transverse direction. The complications due to the presence of mesiodenses were none in 43.5%, diastema in 19.4%, tooth displacement in 17.7%, delayed eruption or impaction in 12.9%, tooth rotation in 4.8%, and dentigerous cyst in 1.7%. Dental cone-beam CT images with 3D provided 3-dimensional perception of mesiodens to the neighboring teeth. This results would be helpful for management of the impacted mesiodens.

  20. Performance evaluation of the backprojection filtered (BPF) algorithm in circular fan-beam and cone-beam CT

    International Nuclear Information System (INIS)

    Li Liang; Chen Zhiqiang; Zhang Li; Kang Kejun

    2006-01-01

    In this article we introduce an exact backprojection filtered (BPF) type reconstruction algorithm for cone-beam scans based on Zou and Pan's work. The algorithm can reconstruct images using only the projection data passing through the parallel PI-line segments in reduced scans. Computer simulations and practical experiments are carried out to evaluate this algorithm. the BPF algorithm has a higher computational efficiency than the famous FDK algorithm. the BPF algorithm is evaluated using the practical CT projection data on a 450 keV X-ray CT system with a flat-panel detector (FPD). From the practical experiments, we get the spatial resolution of this CT system. The algorithm could achieve the spatial resolution of 2.4 lp/mm and satisfies the practical applications in industrial CT inspection. (authors)

  1. The noise power spectrum in CT with direct fan beam reconstruction

    International Nuclear Information System (INIS)

    Baek, Jongduk; Pelc, Norbert J.

    2010-01-01

    The noise power spectrum (NPS) is a useful metric for understanding the noise content in images. To examine some unique properties of the NPS of fan beam CT, the authors derived an analytical expression for the NPS of fan beam CT and validated it with computer simulations. The nonstationary noise behavior of fan beam CT was examined by analyzing local regions and the entire field-of-view (FOV). This was performed for cases with uniform as well as nonuniform noise across the detector cells and across views. The simulated NPS from the entire FOV and local regions showed good agreement with the analytically derived NPS. The analysis shows that whereas the NPS of a large FOV in parallel beam CT (using a ramp filter) is proportional to frequency, the NPS with direct fan beam FBP reconstruction shows a high frequency roll off. Even in small regions, the fan beam NPS can show a sharp transition (discontinuity) at high frequencies. These effects are due to the variable magnification and therefore are more pronounced as the fan angle increases. For cases with nonuniform noise, the NPS can show the directional dependence and additional effects.

  2. Image characteristics of cone beam computed tomography using a CT performance phantom

    International Nuclear Information System (INIS)

    Han, Choong Wan; Kim, Gyu Tae; Choi, Yong Suk; Hwang, Eui Hwan

    2007-01-01

    To evaluate the characteristics of (widely used) cone beam computed tomography (CBCT) images. Images were obtained with CT performance phantoms (The American Association of Physicists in Medicine; AAPM). CT phantom as the destination by using PSR 9000N TM dental CT system (Asahi Roentgen Ind. Co., Ltd., Japan) and i-CAT CBCT (Imaging Science International Inc., USA) that have different kinds of detectors and field of view, and compared these images with the CT number for linear attenuation, contrast resolution, and spatial resolution. CT number of both PSR 9000N TM dental CT system and i-CAT CBCT did not conform to the base value of CT performance phantom. The contrast of i-CAT CBCT is higher than that of PSR 9000N T M dental CT system. Both contrasts were increased according to thickness of cross section. Spatial resolution and shapes of reappearance was possible up to 0.6 mm in PSR 9000N T M dental CT system and up to 1.0 mm in i-CAT CBCT. Low contrast resolution in region of low contrast sensitivity revealed low level at PSR 9000N T M dental CT system and i-CAT CBCT. CBCT images revealed higher spatial resolution, however, contrast resolution in region of low contrast sensitivity was the inferiority of image characteristics

  3. Cone beam CT findings of retromolar canals: Report of cases and literature review

    Energy Technology Data Exchange (ETDEWEB)

    Han, Sang Sun [Dept. of Dental Hygiene, Eulji University, Seongnam (Korea, Republic of); Park, Chang Seo [Dept. of Oral and Maxillofacial Radiology, College of Dentistry, Yonsei University, Seoul (Korea, Republic of)

    2013-12-15

    A retromolar canal is an anatomical variation in the mandible. As it includes the neurovascular bundle, local anesthetic insufficiency can occur, and an injury of the retromolar canal during dental surgery in the mandible may result in excessive bleeding, paresthesia, and traumatic neuroma. Using imaging analysis software, we evaluated the cone-beam computed tomography (CT) images of two Korean patients who presented with retromolar canals. Retromolar canals were detectable on the sagittal and cross-sectional images of cone-beam CT, but not on the panoramic radiographs of the patients. Therefore, the clinician should pay particular attention to the identification of retromolar canals by preoperative radiographic examination, and additional cone beam CT scanning would be recommended.

  4. Pre-surgical treatment planning of maxillary canine impactions using panoramic vs cone beam CT imaging.

    Science.gov (United States)

    Alqerban, A; Hedesiu, M; Baciut, M; Nackaerts, O; Jacobs, R; Fieuws, S; Willems, G

    2013-01-01

    The aim of this prospective study was to compare the impact of using two-dimensional (2D) panoramic radiographs and three-dimensional (3D) cone beam CT for the surgical treatment planning of impacted maxillary canines. This study consisted of 32 subjects (19 females, 13 males) with a mean age of 25 years, referred for surgical intervention of 39 maxillary impacted canines. Initial 2D panoramic radiography was available, and 3D cone beam CT imaging was obtained upon clinical indication. Both 2D and 3D pre-operative radiographic diagnostic sets were subsequently analysed by six observers. Perioperative evaluations were conducted by the treating surgeon. McNemar tests, hierarchical logistic regression and linear mixed models were used to explore the differences in evaluations between imaging modalities. Significantly higher confidence levels were observed for 3D image-based treatment plans than for 2D image-based plans (p panoramic and cone beam CT images.

  5. Cone beam CT findings of retromolar canals: Report of cases and literature review

    International Nuclear Information System (INIS)

    Han, Sang Sun; Park, Chang Seo

    2013-01-01

    A retromolar canal is an anatomical variation in the mandible. As it includes the neurovascular bundle, local anesthetic insufficiency can occur, and an injury of the retromolar canal during dental surgery in the mandible may result in excessive bleeding, paresthesia, and traumatic neuroma. Using imaging analysis software, we evaluated the cone-beam computed tomography (CT) images of two Korean patients who presented with retromolar canals. Retromolar canals were detectable on the sagittal and cross-sectional images of cone-beam CT, but not on the panoramic radiographs of the patients. Therefore, the clinician should pay particular attention to the identification of retromolar canals by preoperative radiographic examination, and additional cone beam CT scanning would be recommended.

  6. An analytical simulation technique for cone-beam CT and pinhole SPECT

    International Nuclear Information System (INIS)

    Zhang Xuezhu; Qi Yujin

    2011-01-01

    This study was aimed at developing an efficient simulation technique with an ordinary PC. The work involved derivation of mathematical operators, analytic phantom generations, and effective analytical projectors developing for cone-beam CT and pinhole SPECT imaging. The computer simulations based on the analytical projectors were developed by ray-tracing method for cone-beam CT and voxel-driven method for pinhole SPECT of degrading blurring. The 3D Shepp-Logan, Jaszczak and Defrise phantoms were used for simulation evaluations and image reconstructions. The reconstructed phantom images were of good accuracy with the phantoms. The results showed that the analytical simulation technique is an efficient tool for studying cone-beam CT and pinhole SPECT imaging. (authors)

  7. Investigation of bulk electron densities for dose calculations on cone-beam CT images

    International Nuclear Information System (INIS)

    Lambert, J.; Parker, J.; Gupta, S.; Hatton, J.; Tang, C.; Capp, A.; Denham, J.W.; Wright, P.

    2010-01-01

    Full text: If cone-beam CT images are to be used for dose calculations, then the images must be able to provide accurate electron density information. Twelve patients underwent twice weekly cone-beam CT scans in addition to the planning CT scan. A standardised 5-field treatment plan was applied to 169 of the CBCT images. Doses were calculated using the original electron density values in the CBCT and with bulk electron densities applied. Bone was assigned a density of 288 HU, and all other tissue was assigned to be water equivalent (0 HU). The doses were compared to the dose calculated on the original planning CT image. Using the original HU values in the cone-beam images, the average dose del i vered by the plans from all 12 patients was I. I % lower than the intended 200 cOy delivered on the original CT plans (standard devia tion 0.7%, maximum difference -2.93%). When bulk electron densities were applied to the cone-beam images, the average dose was 0.3% lower than the original CT plans (standard deviation 0.8%, maximum difference -2.22%). Compared to using the original HU values, applying bulk electron densities to the CBCT images improved the dose calculations by almost I %. Some variation due to natural changes in anatomy should be expected. The application of bulk elec tron densities to cone beam CT images has the potential to improve the accuracy of dose calculations due to inaccurate H U values. Acknowledgements This work was partially funded by Cancer Council NSW Grant Number RG 07-06.

  8. Clinical experience with cone-beam CT navigation for tumor ablation.

    Science.gov (United States)

    Abi-Jaoudeh, Nadine; Venkatesan, Aradhana M; Van der Sterren, William; Radaelli, Alessandro; Carelsen, Bart; Wood, Bradford J

    2015-02-01

    To describe clinical use and potential benefits of cone-beam computed tomography (CT) navigation to perform image-guided percutaneous tumor ablation. All ablations performed between February 2011 and February 2013 using cone-beam CT navigation were included. There were 16 patients who underwent 20 ablations for 29 lesions. Cone-beam CT ablation planning capabilities include multimodality image fusion and tumor segmentation for visualization, depiction of the predicted ablation zones for intraprocedural planning, and segmentation of the ablated area for immediate verification after treatment. Number and purpose of cone-beam CT scans were examined. The initial ablation plan, defined as number of probes and duration of energy delivery, was recorded for the 20 of the 29 lesions ablated. Technical success and local recurrences were recorded. Primary and secondary effectiveness rates were calculated. Image fusion was used for 16 lesions, and intraprocedural ultrasound was used for 4 lesions. Of the 20 ablations, where the ablation plans were recorded, there was no deviation from the plan in 14 ablations. In the remaining 6 ablations, iterative planning was needed for complete tumor coverage. An average of 8.7 cone-beam CT scans ± 3.2 were performed per procedure, including 1.3 ± 0.5 for tumor segmentation and planning, 1.7 ± 0.7 for probe position confirmation, and 3.9 ± 2 to ensure complete coverage. Mean follow-up time was 18.6 months ± 6.5. Ablations for 28 of 29 lesions were technically successful (96.5%). Of ablations performed with curative intent, technical effectiveness at 1 month was 25 of 26 lesions (96.1%) and 22 of 26 lesions (84.6%) at last follow-up. Local tumor progression was observed in 11.5% (3 of 26 lesions). Cone-beam CT navigation may add information to assist and improve ablation guidance and monitoring. Published by Elsevier Inc.

  9. Protocol of image guided off-line using cone beam CT megavoltage; Protocolo de imagen guiada off-line mediante Cone Beam CT de megavoltaje

    Energy Technology Data Exchange (ETDEWEB)

    Garcia Ruiz-Zorrilla, J.; Fernandez Leton, J. P.; Perez Moreno, J. M.; Zucca Aparicio, D.; Minambres Moro, A.

    2013-07-01

    The goal of image guided protocols offline is to reduce systematic errors in positioning of the patient in the treatment unit, being more important than the random errors, since the systematic have one contribution in the margin of the CTV to the PTV. This paper proposes a protocol for image guided offline with the different actions to take with their threshold values evaluated previously by anatomic location in a sample of 474 patients and 4821Cone beam Megavoltaje CT (CBCT). (Author)

  10. Point spread function modeling and image restoration for cone-beam CT

    International Nuclear Information System (INIS)

    Zhang Hua; Shi Yikai; Huang Kuidong; Xu Zhe

    2015-01-01

    X-ray cone-beam computed tomography (CT) has such notable features as high efficiency and precision, and is widely used in the fields of medical imaging and industrial non-destructive testing, but the inherent imaging degradation reduces the quality of CT images. Aimed at the problems of projection image degradation and restoration in cone-beam CT, a point spread function (PSF) modeling method is proposed first. The general PSF model of cone-beam CT is established, and based on it, the PSF under arbitrary scanning conditions can be calculated directly for projection image restoration without the additional measurement, which greatly improved the application convenience of cone-beam CT. Secondly, a projection image restoration algorithm based on pre-filtering and pre-segmentation is proposed, which can make the edge contours in projection images and slice images clearer after restoration, and control the noise in the equivalent level to the original images. Finally, the experiments verified the feasibility and effectiveness of the proposed methods. (authors)

  11. Region-of-interest reconstruction for a cone-beam dental CT with a circular trajectory

    International Nuclear Information System (INIS)

    Hu, Zhanli; Zou, Jing; Gui, Jianbao; Zheng, Hairong; Xia, Dan

    2013-01-01

    Dental CT is the most appropriate and accurate device for preoperative evaluation of dental implantation. It can demonstrate the quantity of bone in three dimensions (3D), the location of important adjacent anatomic structures and the quality of available bone with minimal geometric distortion. Nevertheless, with the rapid increase of dental CT examinations, we are facing the problem of dose reduction without loss of image quality. In this work, backprojection-filtration (BPF) and Feldkamp–Davis–Kress (FDK) algorithm was applied to reconstruct the 3D full image and region-of-interest (ROI) image from complete and truncated circular cone-beam data respectively by computer-simulation. In addition, the BPF algorithm was evaluated based on the 3D ROI-image reconstruction from real data, which was acquired from our developed circular cone-beam prototype dental CT system. The results demonstrated that the ROI-image quality reconstructed from truncated data using the BPF algorithm was comparable to that reconstructed from complete data. The FDK algorithm, however, created artifacts while reconstructing ROI-image. Thus it can be seen, for circular cone-beam dental CT, reducing scanning angular range of the BPF algorithm used for ROI-image reconstruction are helpful for reducing the radiation dose and scanning time. Finally, an analytical method was developed for estimation of the ROI projection area on the detector before CT scanning, which would help doctors to roughly estimate the total radiation dose before the CT examination. -- Highlights: ► BPF algorithm was applied by using dental CT for the first time. ► A method was developed for estimation of projection region before CT scanning. ► Roughly predict the total radiation dose before CT scans. ► Potential reduce imaging radiation dose, scatter, and scanning time

  12. 3D Analytic Cone-Beam Reconstruction for Multiaxial CT Acquisitions

    Science.gov (United States)

    Yin, Zhye; De Man, Bruno; Pack, Jed

    2009-01-01

    A conventional 3rd generation Computed Tomography (CT) system with a single circular source trajectory is limited in terms of longitudinal scan coverage since extending the scan coverage beyond 40 mm results in significant cone-beam artifacts. A multiaxial CT acquisition is achieved by combining multiple sequential 3rd generation axial scans or by performing a single axial multisource CT scan with multiple longitudinally offset sources. Data from multiple axial scans or multiple sources provide complementary information. For full-scan acquisitions, we present a window-based 3D analytic cone-beam reconstruction algorithm by tessellating data from neighboring axial datasets. We also show that multi-axial CT acquisition can extend the axial scan coverage while minimizing cone-beam artifacts. For half-scan acquisitions, one cannot take advantage of conjugate rays. We propose a cone-angle dependent weighting approach to combine multi-axial half-scan data. We compute the relative contribution from each axial dataset to each voxel based on the X-ray beam collimation, the respective cone-angles, and the spacing between the axial scans. We present numerical experiments to demonstrate that the proposed techniques successfully reduce cone-beam artifacts at very large volumetric coverage. PMID:19730750

  13. Pulmonary Artery Imaging in Patients with Chronic Thromboembolic Pulmonary Hypertension: Comparison of Cone-Beam CT and 64-Row Multidetector CT.

    Science.gov (United States)

    Hinrichs, Jan B; von Falck, Christian; Hoeper, Marius M; Olsson, Karen M; Wacker, Frank K; Meyer, Bernhard C; Renne, Julius

    2016-03-01

    To compare the depiction of pulmonary arteries in pulmonary arterial catheter-based contrast-enhanced cone-beam CT with peripheral intravenous contrast-enhanced multidetector CT in patients with suspected chronic thromboembolic pulmonary hypertension. In 20 patients (15 men and 5 women, 63.4 y ± 16.3), cone-beam CT using a catheter placed in the main pulmonary artery and 64-row multidetector CT using an appropriate venous access were performed. Contrast enhancement was measured in the main pulmonary artery, the right and left pulmonary arteries, and the left atrium. The amount of peripheral vessel conspicuity adjacent to the pleural surface (distance from vessel-to pleura) was measured. Two readers (R1, R2) independently evaluated the pulmonary arteries for image quality and pathologic findings in both modalities. Contrast density was higher in the main pulmonary artery and right and left pulmonary arteries (P cone-beam CT. The smallest distance between clearly delineated vessels and the pleura was significantly lower on cone-beam CT images (P cone-beam CT (κ = 0.79) and multidetector CT (κ = 0.78), whereas intermodality agreement was moderate (R1, κ = 0.60; R2, κ = 0.59). Both readers detected more weblike stenoses with cone-beam CT (76; 22%) compared with multidetector CT (25; 7%). Cone-beam CT shows improved contrast between pulmonary arteries and the left atrium and allows a more detailed depiction of the pulmonary arteries. Copyright © 2016 SIR. Published by Elsevier Inc. All rights reserved.

  14. SU-E-J-99: Reconstruction of Cone Beam CT Image Using Volumetric Modulated Arc Therapy Exit Beams

    Energy Technology Data Exchange (ETDEWEB)

    Jeong, K; Goddard, L; Savacool, M; Mynampati, D; Godoy Scripes, P; Tome' , W [Montefiore Medical Center, Bronx, NY (United States); Kuo, H; Basavatia, A; Hong, L; Yaparpalvi, R; Kalnicki, S [Montefiore Medical Center/Albert Einstein College of Medicine, Bronx, NY (United States)

    2014-06-01

    Purpose: To test the possibility of obtaining an image of the treated volume during volumetric modulated arc therapy (VMAT) with exit beams. Method: Using a Varian Clinac 21EX and MVCT detector the following three sets of detector projection data were obtained for cone beam CT reconstruction with and without a Catphan 504 phantom. 1) 72 projection images from 20 × 16 cm{sup 2} open beam with 3 MUs, 2) 72 projection images from 20 × 16 cm{sup 2} MLC closed beam with 14 MUs. 3) 137 projection images from a test RapicArc QA plan. All projection images were obtained in ‘integrated image’ mode. We used OSCaR code to reconstruct the cone beam CT images. No attempts were made to reduce scatter or artifacts. Results: With projection set 1) we obtained a good quality MV CBCT image by optimizing the reconstruction parameters. Using projection set 2) we were not able to obtain a CBCT image of the phantom, which was determined to be due to the variation of interleaf leakage with gantry angle. From projection set 3), we were able to obtain a weak but meaningful signal in the image, especially in the target area where open beam signals were dominant. This finding suggests that one might be able to acquire CBCT images with rough body shape and some details inside the irradiated target area. Conclusion: Obtaining patient images using the VMAT exit beam is challenging but possible. We were able to determine sources of image degradation such as gantry angle dependent interleaf leakage and beams with a large scatter component. We are actively working on improving image quality.

  15. Feasibility of a Modified Cone-Beam CT Rotation Trajectory to Improve Liver Periphery Visualization during Transarterial Chemoembolization.

    Science.gov (United States)

    Schernthaner, Rüdiger E; Chapiro, Julius; Sahu, Sonia; Withagen, Paul; Duran, Rafael; Sohn, Jae Ho; Radaelli, Alessandro; van der Bom, Imramsjah Martin; Geschwind, Jean-François H; Lin, MingDe

    2015-12-01

    To compare liver coverage and tumor detectability by using preprocedural magnetic resonance (MR) images as a reference, as well as radiation exposure of cone-beam computed tomography (CT) with different rotational trajectories. Fifteen patients (nine men and six women; mean age ± standard deviation, 65 years ± 5) with primary or secondary liver cancer were retrospectively included in this institutional review board-approved study. A modified cone-beam CT protocol was used in which the C-arm rotates from +55° to -185° (open arc cone-beam CT) instead of -120° to +120° (closed arc cone-beam CT). Each patient underwent two sessions of transarterial chemoembolization between February 2013 and March 2014 with closed arc and open arc cone-beam CT (during the first and second transarterial chemoembolization sessions, respectively, as part of the institutional transarterial chemoembolization protocol). For each cone-beam CT examination, liver volume and tumor detectability were assessed by using MR images as the reference. Radiation exposure was compared by means of a phantom study. For statistical analysis, paired t tests and a Wilcoxon signed rank test were performed. Mean liver volume imaged was 1695 cm(3) ± 542 and 1857 cm(3) ± 571 at closed arc and open arc cone-beam CT, respectively. The coverage of open arc cone-beam CT was significantly higher compared with closed arc cone-beam CT (97% vs 86% of the MR imaging liver volume, P = .002). In eight patients (53%), tumors were partially or completely outside the closed arc cone-beam CT field of view. All tumors were within the open arc cone-beam CT field of view. The open arc cone-beam CT radiation exposure by means of weighted CT index was slightly lower compared with that of closed arc cone-beam CT (-5.1%). Open arc cone-beam CT allowed for a significantly improved intraprocedural depiction of peripheral hepatic tumors while achieving a slight radiation exposure reduction.

  16. Exact cone beam CT with a spiral scan

    International Nuclear Information System (INIS)

    Tam, K.C.; Samarasekera, S.; Sauer, F.

    1998-01-01

    A method is developed which makes it possible to scan and reconstruct an object with cone beam x-rays in a spiral scan path with area detectors much shorter than the length of the object. The method is mathematically exact. If only a region of interest of the object is to be imaged, a top circle scan at the top level of the region of interest and a bottom circle scan at the bottom level of the region of interest are added. The height of the detector is required to cover only the distance between adjacent turns in the spiral projected at the detector. To reconstruct the object, the Radon transform for each plane intersecting the object is computed from the totality of the cone beam data. This is achieved by suitably combining the cone beam data taken at different source positions on the scan path; the angular range of the cone beam data required at each source position can be determined easily with a mask which is the spiral scan path projected on the detector from the current source position. The spiral scan algorithm has been successfully validated with simulated cone beam data. (author)

  17. Comparison between beam-stop and beam-hole array scatter correction techniques for industrial X-ray cone-beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Schoerner, K., E-mail: karsten.schoerner.ext@siemens.co [Corporate Technology, Siemens AG, 81739 Muenchen (Germany); Physik-Department, Technische Universitaet Muenchen, 85748 Garching (Germany); Goldammer, M.; Stephan, J. [Corporate Technology, Siemens AG, 81739 Muenchen (Germany)

    2011-02-01

    Research highlights: {yields} We propose a scatter correction method employing a beam-hole array. {yields} Beam-hole and beam-stop array techniques are compared in respect of geometric and scattering properties. {yields} The beam-hole array method reduces overall scattering compared to a beam-stop array. {yields} Application of the beam-hole array method is successfully demonstrated for a CT of ceramic specimen. -- Abstract: In industrial X-ray cone-beam computed tomography, the inspection of large-scale samples is important because of increasing demands on their quality and long-term mechanical resilience. Large-scale samples, for example made of aluminum or iron, are strongly scattering X-rays. Scattered radiation leads to artifacts such as cupping, streaks, and a reduction in contrast in the reconstructed CT-volume. We propose a scatter correction method based on sampling primary signals by employing a beam-hole array (BHA). In this indirect method, a scatter estimate is calculated by subtraction of the sampled primary signal from the total signal, the latter taken from an image where the BHA is absent. This technique is considered complementary to the better known beam-stop array (BSA) method. The two scatter estimation methods are compared here with respect to geometric effects, scatter-to-total ratio and practicability. Scatter estimation with the BHA method yields more accurate scatter estimates in off-centered regions, and a lower scatter-to-total ratio in critical image regions where the primary signal is very low. Scatter correction with the proposed BHA method is then applied to a ceramic specimen from power generation technologies. In the reconstructed CT volume, cupping almost completely vanishes and contrast is enhanced significantly.

  18. Investigation of respiration induced intra- and inter-fractional tumour motion using a standard Cone Beam CT

    DEFF Research Database (Denmark)

    Gottlieb, Karina Lindberg; Hansen, Christian R; Hansen, Olfred

    2010-01-01

    To investigate whether a standard Cone beam CT (CBCT) scan can be used to determined the intra- and inter-fractional tumour motion for lung tumours that have infiltrated the mediastinum.......To investigate whether a standard Cone beam CT (CBCT) scan can be used to determined the intra- and inter-fractional tumour motion for lung tumours that have infiltrated the mediastinum....

  19. Beam characteristics and radiation output of a kilovoltage cone-beam CT

    Science.gov (United States)

    Ding, George X.; Coffey, Charles W.

    2010-09-01

    This study presents beam characteristics of five recently available x-ray beams produced by an on-board imager (OBI 1.4) for acquiring kilovoltage cone-beam computed tomography (kV-CBCT) and investigates suitable methods for the beam radiation output determination resulting from an image acquisition. Both are essential for commissioning an x-ray beam in a radiotherapy treatment planning system. The BEAM/DOSXYZnrc Monte Carlo codes were used in the investigation. The simulated beam data were benchmarked against measurements. Three different commercially available plastic phantom materials are investigated as liquid water substitutes in the beam radiation output determination. Ionization chambers are used for the measurements. Five kV-CBCT beam characteristics including photon fluence, average beam energy and photon spectra are generated from Monte Carlo simulations. The Monte Carlo calculated dose profiles are validated by measurements. The fluence of kV-CBCT beams is strongly dependent on the geometry of added filters as well as X and Y beam collimations. The potential errors of determining the beam output of a kV-CBCT beam in Solid Water and PMMA phantoms may approach 8% and 20%, respectively, for use in a conventional treatment planning system, whereas using the Plastic Water low-energy range (PW-LR) phantom results in errors within 2%. The Monte Carlo simulation is essential in providing the parameters of an x-ray beam which are needed for the commissioning of a kV-CBCT beam in a radiotherapy treatment planning system. The PW-LR phantom is a suitable liquid water substitute in the beam output determination resulting from a kV-CBCT acquisition.

  20. [Digital volume tomography : Dedicated scanner and cone beam CT with C‑arm systems].

    Science.gov (United States)

    Fiebich, M; Weber, D

    2018-03-01

    Digital volume tomography (DVT) and cone-beam computed tomography (CT) with C‑arm systems have become established three-dimensional imaging systems as an alternative to CT in some application areas. The technology of the systems is well developed so that they have become a competing method to CT imaging in terms of image quality and radiation exposure. An advantage is the better spatial resolution, preferably with dedicated scanner systems, especially in the z direction. The radiation exposure of CT, cone beam CT and DVT are comparable, if the exposure parameter in CT imaging can be adjusted to the lower exposure levels. Advantages of these systems are that they can be used for imaging in a better workflow or to acquire images under conditions not possible in CT, e. g. imaging under stress in orthopedics or to take images in the corona technique with a horizontal gantry in cone-beam CT mammography PRACTICAL RECOMMENDATIONS: The use of three-dimensional imaging is becoming more frequent and will replace planar radiography in additional clinical situations. The three-dimensional imaging without superpositioning of structures has advantages in the visibility of structures and the spatial relation to other organs and structures. In guidelines and recommendations, the number of recommendations given for the use of three-dimensional imaging is increasing. This leads to a small increase in the radiation exposure of patients, a trend which is reflected in the annual reports of the Federal Office for Radiation Protection.

  1. Feasibility of contrast-enhanced cone-beam CT for target localization and treatment monitoring

    International Nuclear Information System (INIS)

    Rodal, Jan; Sovik, Aste; Skogmo, Hege Kippenes; Knudtsen, Ingerid Skjei; Malinen, Eirik

    2010-01-01

    A dog with a spontaneous maxillary tumour was given 40 Gy of fractionated radiotherapy. At five out of 10 fractions cone-beam CT (CBCT) imaging before and after administration of an iodinated contrast agent were performed. Contrast enhancement maps were overlaid on the pre-contrast CBCT images. The tumour was clearly visualized in the images thus produced.

  2. Comparison of percutaneous radiologic gastrostomy by using cone beam CT and endoscopic gastrostomy

    International Nuclear Information System (INIS)

    Jung, Hyun Nyeong; Han, Young Min; Jin, Gong Yong; Choi, Eun Jeong; Song, Ji Soo

    2014-01-01

    To compare the effectiveness of percutaneous radiologic gastrostomy (PRG) by using cone beam CT and percutaneous endoscopic gastrostomy (PEG). This study retrospectively reviewed 129 patients who underwent PRG (n = 53) and PEG (n = 76) over a 2-years period. The C-arm cone beam CT images were obtained from all PRG patients before the procedure in order to decide the safest accessing routes. The parameters including technical success rates, complication rates and tube migration rates were all analyzed according to statistical methods. The success rate of tube placement was higher in PRG than in PEG (100% to 93%, p = 0.08). Minor complications occurred in 5 patients of the PRG group (10%; 5/53, 3 wound infection, 2 blood oozing), and occurred in 6 patients of PEG group (7.9%; 6/76, 5 wound infection, 1 esophageal ulcer). Major complications occurred only in 5 patients of PEG group (6.6%; 5/76, 1 panperitonitis, 4 buried bumper syndrome). There were no statistical differences of minor and major complication rates in the two groups (respectively, p = 0.759, p = 0.078). Tube migration rate was lower in PRG than PEG group (7.5% vs. 38.2%, p < 0.005). PRG using cone beam CT is the effective and safe method, the cone beam CT provides the safest accessing route during gastrostomy. Less tube migration occurs in the PRG than in PEG.

  3. Cone-beam CT in paediatric dentistry: DIMITRA project position statement.

    Science.gov (United States)

    Oenning, Anne Caroline; Jacobs, Reinhilde; Pauwels, Ruben; Stratis, Andreas; Hedesiu, Mihaela; Salmon, Benjamin

    2018-03-01

    DIMITRA (dentomaxillofacial paediatric imaging: an investigation towards low-dose radiation induced risks) is a European multicenter and multidisciplinary project focused on optimizing cone-beam CT exposures for children and adolescents. With increasing use of cone-beam CT for dentomaxillofacial diagnostics, concern arises regarding radiation risks associated with this imaging modality, especially for children. Research evidence concerning cone-beam CT indications in children remains limited, while reports mention inconsistent recommendations for dose reduction. Furthermore, there is no paper using the combined and integrated information on the required indication-oriented image quality and the related patient dose levels. In this paper, therefore, the authors initiate an integrated approach based on current evidence regarding image quality and dose, together with the expertise of DIMITRA's members searching for a state of the art. The aim of this DIMITRA position statement is to provide indication-oriented and patient-specific recommendations regarding the main cone-beam CT applications in the pediatric field. The authors will review this position statement document when results regarding multidisciplinary approaches evolve, in a period of 5 years or earlier.

  4. Measurements on 3D models of human skulls derived from two different cone beam CT scanners

    NARCIS (Netherlands)

    van Vlijmen, Olivier J. C.; Rangel, Frits A.; Bergé, Stefaan J.; Bronkhorst, Ewald M.; Becking, Alfred G.; Kuijpers-Jagtman, Anne Marie

    2011-01-01

    The aims of this study were to compare measurements on three-dimensional (3D) models of human skulls derived from two different cone beam CT scanners (CBCT) and to evaluate if the used hardware can influence the performed measurements. CBCT scans of 40 dry human skulls with both the i-CAT and the

  5. Segmentation of the Mandibular Canal in Cone-Beam CT Data

    NARCIS (Netherlands)

    Kroon, Dirk-Jan

    2011-01-01

    Accurate information about the location of the mandibular canal is essential in case of dental implant surgery. The goal of our research is to find an automatic method which can segment the mandibular canal in Cone-beam CT (CBCT). Mandibular canal segmentation methods in literature using a priori

  6. Evaluation of tilted cone-beam CT orbits in the development of a dedicated hybrid mammotomograph

    Energy Technology Data Exchange (ETDEWEB)

    Madhav, P; Crotty, D J; Tornai, M P [Department of Radiology, Duke University Medical Center, Durham, NC 27710 (United States); McKinley, R L [Zumatek Incorporated, Chapel Hill, NC 27519 (United States)], E-mail: priti.madhav@duke.edu

    2009-06-21

    A compact dedicated 3D breast SPECT-CT (mammotomography) system is currently under development. In its initial prototype, the cone-beam CT sub-system is restricted to a fixed-tilt circular rotation around the patient's pendant breast. This study evaluated stationary-tilt angles for the CT sub-system that will enable maximal volumetric sampling and viewing of the breast and chest wall. Images of geometric/anthropomorphic phantoms were acquired using various fixed-tilt circular and 3D sinusoidal trajectories. The iteratively reconstructed images showed more distortion and attenuation coefficient inaccuracy from tilted cone-beam orbits than from the complex trajectory. Additionally, line profiles illustrated cupping artifacts in planes distal to the central plane of the tilted cone-beam, otherwise not apparent for images acquired with complex trajectories. This indicates that undersampled cone-beam data may be an additional cause of cupping artifacts. High-frequency objects could be distinguished for all trajectories, but their shapes and locations were corrupted by out-of-plane frequency information. Although more acrylic balls were visualized with a fixed-tilt and nearly flat cone-beam at the posterior of the breast, 3D complex trajectories have less distortion and more complete sampling throughout the reconstruction volume. While complex trajectories would ideally be preferred, negatively fixed-tilt source-detector configuration demonstrates minimally distorted patient images.

  7. Evaluation of tilted cone-beam CT orbits in the development of a dedicated hybrid mammotomograph

    International Nuclear Information System (INIS)

    Madhav, P; Crotty, D J; Tornai, M P; McKinley, R L

    2009-01-01

    A compact dedicated 3D breast SPECT-CT (mammotomography) system is currently under development. In its initial prototype, the cone-beam CT sub-system is restricted to a fixed-tilt circular rotation around the patient's pendant breast. This study evaluated stationary-tilt angles for the CT sub-system that will enable maximal volumetric sampling and viewing of the breast and chest wall. Images of geometric/anthropomorphic phantoms were acquired using various fixed-tilt circular and 3D sinusoidal trajectories. The iteratively reconstructed images showed more distortion and attenuation coefficient inaccuracy from tilted cone-beam orbits than from the complex trajectory. Additionally, line profiles illustrated cupping artifacts in planes distal to the central plane of the tilted cone-beam, otherwise not apparent for images acquired with complex trajectories. This indicates that undersampled cone-beam data may be an additional cause of cupping artifacts. High-frequency objects could be distinguished for all trajectories, but their shapes and locations were corrupted by out-of-plane frequency information. Although more acrylic balls were visualized with a fixed-tilt and nearly flat cone-beam at the posterior of the breast, 3D complex trajectories have less distortion and more complete sampling throughout the reconstruction volume. While complex trajectories would ideally be preferred, negatively fixed-tilt source-detector configuration demonstrates minimally distorted patient images.

  8. ASSESSMENT OF EFFECTIVE DOSE FROM CONE BEAM CT IMAGING IN SPECT/CT EXAMINATION IN COMPARISON WITH OTHER MODALITIES.

    Science.gov (United States)

    Tonkopi, Elena; Ross, Andrew A

    2016-12-01

    The aim of this study was to assess radiation dose from the cone beam computed tomography (CBCT) component of single photon emission tomography/computed tomography (SPECT/CT) examinations and to compare it with the radiopharmaceutical related dose as well as dose from multidetector computed tomography (MDCT). Effective dose (ED) from computed tomography (CT) was estimated using dose-length product values and anatomy-specific conversion factors. The contribution from the SPECT component was evaluated using ED per unit administered activity for the radiopharmaceuticals listed in the International Commission on Radiological Protection Publications 80 and 106. With the exception of cardiac studies (0.11 mSv), the CBCT dose (3.96-6.04 mSv) was similar to that from the radiopharmaceutical accounting for 29-56 % of the total ED from the examination. In comparison with MDCT examinations, the CBCT dose was 48 and 42 % lower for abdomen/pelvis and chest/abdomen/pelvis scans, respectively, while in the chest the CBCT scan resulted in higher dose (23 %). Radiation dose from the CT component should be taken into consideration when evaluating total SPECT/CT patient dose. © The Author 2016. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  9. SU-F-I-73: Surface Dose from KV Diagnostic Beams From An On-Board Imager On a Linac Machine Using Different Imaging Techniques and Filters

    Energy Technology Data Exchange (ETDEWEB)

    Ali, I; Hossain, S; Syzek, E; Ahmad, S [University of Oklahoma Health Sciences Center, Department of Radiation Oncology, Oklahoma City, OK (United States)

    2016-06-15

    Purpose: To quantitatively investigate the surface dose deposited in patients imaged with a kV on-board-imager mounted on a radiotherapy machine using different clinical imaging techniques and filters. Methods: A high sensitivity photon diode is used to measure the surface dose on central-axis and at an off-axis-point which is mounted on the top of a phantom setup. The dose is measured for different imaging techniques that include: AP-Pelvis, AP-Head, AP-Abdomen, AP-Thorax, and Extremity. The dose measurements from these imaging techniques are combined with various filtering techniques that include: no-filter (open-field), half-fan bowtie (HF), full-fan bowtie (FF) and Cu-plate filters. The relative surface dose for different imaging and filtering techniques is evaluated quantiatively by the ratio of the dose relative to the Cu-plate filter. Results: The lowest surface dose is deposited with the Cu-plate filter. The highest surface dose deposited results from open fields without filter and it is nearly a factor of 8–30 larger than the corresponding imaging technique with the Cu-plate filter. The AP-Abdomen technique delivers the largest surface dose that is nearly 2.7 times larger than the AP-Head technique. The smallest surface dose is obtained from the Extremity imaging technique. Imaging with bowtie filters decreases the surface dose by nearly 33% in comparison with the open field. The surface doses deposited with the HF or FF-bowtie filters are within few percentages. Image-quality of the radiographic images obtained from the different filtering techniques is similar because the Cu-plate eliminates low-energy photons. The HF- and FF-bowtie filters generate intensity-gradients in the radiographs which affects image-quality in the different imaging technique. Conclusion: Surface dose from kV-imaging decreases significantly with the Cu-plate and bowtie-filters compared to imaging without filters using open-field beams. The use of Cu-plate filter does not affect

  10. An index of beam hardening artifact for two-dimensional cone-beam CT tomographic images: establishment and preliminary evaluation

    Science.gov (United States)

    Yuan, Fusong; Lv, Peijun; Yang, Huifang; Wang, Yong; Sun, Yuchun

    2015-07-01

    Objectives: Based on the pixel gray value measurements, establish a beam-hardening artifacts index of the cone-beam CT tomographic image, and preliminarily evaluate its applicability. Methods: The 5mm-diameter metal ball and resin ball were fixed on the light-cured resin base plate respectively, while four vitro molars were fixed above and below the ball, on the left and right respectively, which have 10mm distance with the metal ball. Then, cone beam CT was used to scan the fixed base plate twice. The same layer tomographic images were selected from the two data and imported into the Photoshop software. The circle boundary was built through the determination of the center and radius of the circle, according to the artifact-free images section. Grayscale measurement tools were used to measure the internal boundary gray value G0, gray value G1 and G2 of 1mm and 20mm artifacts outside the circular boundary, the length L1 of the arc with artifacts in the circular boundary, the circumference L2. Hardening artifacts index was set A = (G1 / G0) * 0.5 + (G2 / G1) * 0.4 + (L2 / L1) * 0.1. Then, the A values of metal and resin materials were calculated respectively. Results: The A value of cobalt-chromium alloy material is 1, and resin material is 0. Conclusion: The A value reflects comprehensively the three factors of hardening artifacts influencing normal oral tissue image sharpness of cone beam CT. The three factors include relative gray value, the decay rate and range of artifacts.

  11. Feasibility study of image guided radiotherapy for lung tumor using online and offline cone-beam CT setup verification

    International Nuclear Information System (INIS)

    Li Hongsheng; Li Baosheng; Lu Jie; Yin Yong; Yu Ningsha; Chen Yiru

    2009-01-01

    Objective: To investigate the feasibility of online and offline cone-beam CT(CBCT) guided radiotherapy for lung cancer. Methods: Fourteen patients with lung tumor treated by three-dimensional conformal radiotherapy were investigated. Online kV CBCT scan, image registration and setup correction were performed before and immediately after radiotherapy. CBCT online-guided correction data were used to calculate the population-based CTV-PTV margins under the condition of non-correction and correction in every fraction respectively. The numbers of initial images and the population-based CTV-PTV margins after the offline compensation of the system setup error were evaluated with the permission of 0.5 mm and 1.5 mm maximal residue error, respectively. Results: Under the condition of non-correction, the required margins for total error were 5.7 mm, 8.0 mm and 7.8 mm in the left-right (x axis), cranio-caudal (y axis) and anterior-posterior(z axis) directions, respectively. When the tumor was corrected in every fraction, the required margins for intra-fraction error were 2.4 mm, 2.4 mm and 2.3 mm in x,y and z axes, respectively. To correct the systematic setup error, 9 sets of CBCT images for 3.3 mm, 3.7 mm and 3.6 mm PTV margins, and 7 sets of CBCT images for 3.9 mm, 4.3 mm and 4.3 mm PTV margins in x,y and z axes were necessary when 0.5 mm and 1.5 mm maximal residue error were permitted respectively. Conclusions: Both of the online CBCT correction and the offline adaptive correction can markedly reduce the impact of setup error and reduce the required PTV margins accordingly. It is feasible to deliver the online and offline image guided radiation for patients with lung tumor. (authors)

  12. Dynamic bowtie filter for cone-beam/multi-slice CT.

    Directory of Open Access Journals (Sweden)

    Fenglin Liu

    Full Text Available A pre-patient attenuator ("bowtie filter" or "bowtie" is used to modulate an incoming x-ray beam as a function of the angle of the x-ray with respect to a patient to balance the photon flux on a detector array. While the current dynamic bowtie design is focused on fan-beam geometry, in this study we propose a methodology for dynamic bowtie design in multi-slice/cone-beam geometry. The proposed 3D dynamic bowtie is an extension of the 2D prior art. The 3D bowtie consists of a highly attenuating bowtie (HB filled in with heavy liquid and a weakly attenuating bowtie (WB immersed in the liquid of the HB. The HB targets a balanced flux distribution on a detector array when no object is in the field of view (FOV. The WB compensates for an object in the FOV, and hence is a scaled-down version of the object. The WB is rotated and translated in synchrony with the source rotation and patient translation so that the overall flux balance is maintained on the detector array. First, the mathematical models of different scanning modes are established for an elliptical water phantom. Then, a numerical simulation study is performed to compare the performance of the scanning modes in the cases of the water phantom and a patient cross-section without any bowtie and with a dynamic bowtie. The dynamic bowtie can equalize the numbers of detected photons in the case of the water phantom. In practical cases, the dynamic bowtie can effectively reduce the dynamic range of detected signals inside the FOV. Furthermore, the WB can be individualized using a 3D printing technique as the gold standard. We have extended the dynamic bowtie concept from 2D to 3D by using highly attenuating liquid and moving a scale-reduced negative copy of an object being scanned. Our methodology can be applied to reduce radiation dose and facilitate photon-counting detection.

  13. Dynamic Bowtie Filter for Cone-Beam/Multi-Slice CT

    Science.gov (United States)

    Liu, Fenglin; Yang, Qingsong; Cong, Wenxiang; Wang, Ge

    2014-01-01

    A pre-patient attenuator (“bowtie filter” or “bowtie”) is used to modulate an incoming x-ray beam as a function of the angle of the x-ray with respect to a patient to balance the photon flux on a detector array. While the current dynamic bowtie design is focused on fan-beam geometry, in this study we propose a methodology for dynamic bowtie design in multi-slice/cone-beam geometry. The proposed 3D dynamic bowtie is an extension of the 2D prior art. The 3D bowtie consists of a highly attenuating bowtie (HB) filled in with heavy liquid and a weakly attenuating bowtie (WB) immersed in the liquid of the HB. The HB targets a balanced flux distribution on a detector array when no object is in the field of view (FOV). The WB compensates for an object in the FOV, and hence is a scaled-down version of the object. The WB is rotated and translated in synchrony with the source rotation and patient translation so that the overall flux balance is maintained on the detector array. First, the mathematical models of different scanning modes are established for an elliptical water phantom. Then, a numerical simulation study is performed to compare the performance of the scanning modes in the cases of the water phantom and a patient cross-section without any bowtie and with a dynamic bowtie. The dynamic bowtie can equalize the numbers of detected photons in the case of the water phantom. In practical cases, the dynamic bowtie can effectively reduce the dynamic range of detected signals inside the FOV. Furthermore, the WB can be individualized using a 3D printing technique as the gold standard. We have extended the dynamic bowtie concept from 2D to 3D by using highly attenuating liquid and moving a scale-reduced negative copy of an object being scanned. Our methodology can be applied to reduce radiation dose and facilitate photon-counting detection. PMID:25051067

  14. Characterization of CT beams using Compton spectrometry; Caracterização de feixes de TC utilizando Espectrometria Compton

    Energy Technology Data Exchange (ETDEWEB)

    Terini, Ricardo A.; Nerssissian, Denise Y.; Campelo, Maria Carolina S.; Yoshimura, Elisabeth M., E-mail: rterini@if.usp.br [Universidade de São Paulo (LDRFM/USP), SP (Brazil). Lab. de Dosimetria das Radiações e Física Médica

    2017-07-01

    Obtaining the energy spectra of computed tomography (CT) X-ray beams is essential, helping to obtain parameters that characterize beam quality and equipment performance. However, CT photon fluxes are too high to have the spectra measured directly with common photon counting detectors. In this work, a Compton spectrometer was designed, with Al-Pb-Al collimators and shields, as well as a cadmium telluride (CdTe) detector to get the spectrum of CT beams, from the measurement of the spectrum of a beam scattered at 90 deg by a polymethyl-methacrylate (PMMA) rod. A MatLab® computer code was developed, using the Waller-Hartree formalism, to reconstruct the spectrum of the incident beam, from the measured scattered beam spectrum. Tests at IF-USP Laboratory of Radiation Dosimetry and Medical Physics with standard CT beams showed that the reconstructed spectrum is alike the directly measured beam. Shielding influence and scatterer thickness were investigated. The system was tested in measurements on a GE 690 CT scanner, showing practical positioning on the exam table, and alignment with CT lasers refined by scan projection radiography. Spectra obtained with the properly shielded system presented values of half-value layer (HVL) compatible with those measured in QC tests and kVp values with accuracy to evaluate the scanner voltage calibration. (author)

  15. Scatter correction, intermediate view estimation and dose characterization in megavoltage cone-beam CT imaging

    Science.gov (United States)

    Sramek, Benjamin Koerner

    The ability to deliver conformal dose distributions in radiation therapy through intensity modulation and the potential for tumor dose escalation to improve treatment outcome has necessitated an increase in localization accuracy of inter- and intra-fractional patient geometry. Megavoltage cone-beam CT imaging using the treatment beam and onboard electronic portal imaging device is one option currently being studied for implementation in image-guided radiation therapy. However, routine clinical use is predicated upon continued improvements in image quality and patient dose delivered during acquisition. The formal statement of hypothesis for this investigation was that the conformity of planned to delivered dose distributions in image-guided radiation therapy could be further enhanced through the application of kilovoltage scatter correction and intermediate view estimation techniques to megavoltage cone-beam CT imaging, and that normalized dose measurements could be acquired and inter-compared between multiple imaging geometries. The specific aims of this investigation were to: (1) incorporate the Feldkamp, Davis and Kress filtered backprojection algorithm into a program to reconstruct a voxelized linear attenuation coefficient dataset from a set of acquired megavoltage cone-beam CT projections, (2) characterize the effects on megavoltage cone-beam CT image quality resulting from the application of Intermediate View Interpolation and Intermediate View Reprojection techniques to limited-projection datasets, (3) incorporate the Scatter and Primary Estimation from Collimator Shadows (SPECS) algorithm into megavoltage cone-beam CT image reconstruction and determine the set of SPECS parameters which maximize image quality and quantitative accuracy, and (4) evaluate the normalized axial dose distributions received during megavoltage cone-beam CT image acquisition using radiochromic film and thermoluminescent dosimeter measurements in anthropomorphic pelvic and head and

  16. Individualized volume CT dose index determined by cross-sectional area and mean density of the body to achieve uniform image noise of contrast-enhanced pediatric chest CT obtained at variable kV levels and with combined tube current modulation

    Energy Technology Data Exchange (ETDEWEB)

    Goo, Hyun Woo [University of Ulsan College of Medicine, Department of Radiology and Research Institute of Radiology, Asan Medical Center, Seoul (Korea, Republic of)

    2011-07-15

    A practical body-size adaptive protocol providing uniform image noise at various kV levels is not available for pediatric CT. To develop a practical contrast-enhanced pediatric chest CT protocol providing uniform image noise by using an individualized volume CT dose index (CTDIvol) determined by the cross-sectional area and density of the body at variable kV levels and with combined tube current modulation. A total of 137 patients (mean age, 7.6 years) underwent contrast-enhanced pediatric chest CT based on body weight. From the CTDIvol, image noise, and area and mean density of the cross-section at the lung base in the weight-based group, the best fit equation was estimated with a very high correlation coefficient ({gamma}{sup 2} = 0.86, P < 0.001). For the next study, 177 patients (mean age, 7.9 years; the CTDIvol group) underwent contrast-enhanced pediatric chest CT with the CTDIvol determined individually by the best fit equation. CTDIvol values on the dose report after CT scanning, noise differences from the target noise, areas, and mean densities were compared between these two groups. The CTDIvol values (mean{+-}standard deviation, 1.6 {+-} 0.7 mGy) and the noise differences from the target noise (1.1 {+-} 0.9 HU) of the CTDIvol group were significantly lower than those of the weight-based group (2.0 {+-} 1.0 mGy, 1.8 {+-} 1.4 HU) (P < 0.001). In contrast, no statistically significant difference was found in area (317.0 {+-} 136.8 cm{sup 2} vs. 326.3 {+-} 124.8 cm{sup 2}), mean density (-212.9 {+-} 53.1 HU vs. -221.1 {+-} 56.3 HU), and image noise (13.8 {+-} 2.3 vs. 13.6 {+-} 1.7 HU) between the weight-based and the CTDIvol groups (P > 0.05). Contrast-enhanced pediatric chest CT with the CTDIvol determined individually by the cross-sectional area and density of the body provides more uniform noise and better dose adaptation to body habitus than does weight-based CT at variable kV levels and with combined tube current modulation. (orig.)

  17. The diagnosis of atherosclerotic aortic ulcer by electron beam CT

    International Nuclear Information System (INIS)

    Zhi Aihua; Dai Ruping; Jiang Shiliang

    2007-01-01

    Objective: To evaluate the clinical value of electron beam computed tomography (EBCT) in the diagnosis of atherosclerotic aortic ulcer. Methods: Sixty-eight consecutive patients (55 men and 13 women, aged 40-85 years, mean 65.12 ± 9.55 years) with atherosclerotic aortic ulcer, who underwent EBCT scans from December 2001 to December 2004, were studied retrospectively. Contrast-enhanced continuous volume scanning (CVS) was performed by Imatron C-150XP EBCT scanner with 6 mm or 3 mm slice thickness and 100 milliseconds acquisition time. The scan was started 18-30 s after the injection of 80-100 ml contrast medium at the rate of 3.5-4.5 ml/s. Results: In sixty-eight patients with atherosclerotic aortic ulcer, 50 patients had acute aortic syndromes, 36 had intramural hematomas, 15 had atherosclerotic aortic aneurysms, 3 had aortic dissections. 46 patients with progresive ulcer usually had acute aortic syndrome while 22 patients with stable ulcer didn't (P<0.01). Atherosclerotic aortic ulcer was seen more frequently in the aorta arch than other portions of the aorta (P<0.01). Conclusion: EBCT is a very useful tool for the detection and follow-up of atherosclerotic aortic ulcer. (authors)

  18. A virtual source model for Kilo-voltage cone beam CT: Source characteristics and model validation

    International Nuclear Information System (INIS)

    Spezi, E.; Volken, W.; Frei, D.; Fix, M. K.

    2011-01-01

    Purpose: The purpose of this investigation was to study the source characteristics of a clinical kilo-voltage cone beam CT unit and to develop and validate a virtual source model that could be used for treatment planning purposes. Methods: We used a previously commissioned full Monte Carlo model and new bespoke software to study the source characteristics of a clinical kilo-voltage cone beam CT (CBCT) unit. We identified the main particle sources, their spatial, energy and angular distribution for all the image acquisition presets currently used in our clinical practice. This includes a combination of two energies (100 and 120 kVp), two filters (neutral and bowtie), and eight different x-ray beam apertures. We subsequently built a virtual source model which we validated against full Monte Carlo calculations. Results: We found that the radiation output of the clinical kilo-voltage cone beam CT unit investigated in this study could be reproduced with a virtual model comprising of two sources (target and filtration cone) or three sources (target, filtration cone and bowtie filter) when additional filtration was used. With this model, we accounted for more than 97% of the photons exiting the unit. Each source in our model was characterised by a origin distribution in both X and Y directions, a fluence map, a single energy spectrum for unfiltered beams and a two dimensional energy spectrum for bowtie filtered beams. The percentage dose difference between full Monte Carlo and virtual source model based dose distributions was well within the statistical uncertainty associated with the calculations ( ± 2%, one standard deviation) in all cases studied. Conclusions: The virtual source that we developed is accurate in calculating the dose delivered from a commercial kilo-voltage cone beam CT unit operating with routine clinical image acquisition settings. Our data have also shown that target, filtration cone, and bowtie filter sources needed to be all included in the model

  19. Cone beam CT evaluation of patient set-up accuracy as a QA tool

    DEFF Research Database (Denmark)

    Nielsen, Morten; Bertelsen, Anders; Westberg, Jonas

    2009-01-01

    Purpose. To quantify by means of cone beam CT the random and systematic uncertainty involved in radiotherapy, and to determine if this information can be used for e.g. technical quality assurance, evaluation of patient immobilization and determination of margins for the treatment planning. Patients...... and lateral directions). In the CC direction, the margin has to be 5 mm for the Thorax patients. The total uncertainty on the patient position grows during the treatment course, especially in the CC direction for patients receiving thoracical irradiation. This may stem from problems in the immobilization...... and methods. Eighty four cancer patients have been cone beam CT scanned at treatment sessions 1, 2, 3, 10 and 20. Translational and rotational errors are analyzed. Results and conclusions. For the first three treatment sessions the mean translational error in the AP direction is 1 mm; this indicates a small...

  20. MR cone-beam CT fusion image overlay for fluoroscopically guided percutaneous biopsies in pediatric patients.

    Science.gov (United States)

    Thakor, Avnesh S; Patel, Premal A; Gu, Richard; Rea, Vanessa; Amaral, Joao; Connolly, Bairbre L

    2016-03-01

    Lesions only visible on magnetic resonance (MR) imaging cannot easily be targeted for image-guided biopsy using ultrasound or X-rays but instead require MR guidance with MR-compatible needles and long procedure times (acquisition of multiple MR sequences). We developed an alternative method for performing these difficult biopsies in a standard interventional suite, by fusing MR with cone-beam CT images. The MR cone-beam CT fusion image is then used as an overlay to guide a biopsy needle to the target area under live fluoroscopic guidance. Advantages of this technique include (i) the ability for it to be performed in a conventional interventional suite, (ii) three-dimensional planning of the needle trajectory using cross-sectional imaging, (iii) real-time fluoroscopic guidance for needle trajectory correction and (iv) targeting within heterogeneous lesions based on MR signal characteristics to maximize the potential biopsy yield.

  1. Assessment of protocols in cone beam CT with symmetric and asymmetric beam using effective dose and P{sub ka}

    Energy Technology Data Exchange (ETDEWEB)

    Batista, W. O.; Linhares de O, M. V. [Instituto Federal da Bahia, Rua Emidio dos Santos s/n, Barbalho, Salvador, 40301015 Bahia (Brazil); Soares, M. R.; Maia, A. F. [Universidade Federal de Sergipe, Departamento de Fisica, Cidade Universitaria Prof. Jose Aloisio de Campos, Marechal Rondon s/n, Jardim Rosa Elze, 49-100000 Sao Cristovao, Sergipe (Brazil); Caldas, L. V. E., E-mail: wilsonottobatista@gmail.com [Instituto de Pesquisas Energeticas e Nucleares / CNEN, Av. Lineu Prestes 2242, Cidade Universitaria, 05508-000 Sao Paulo (Brazil)

    2014-08-15

    The cone beam CT is an emerging technology in dental radiology with significant differences the point of view of design technology between the various manufacturers on the world market. This study aims to evaluate and compare protocols with similar purposes in a cone beam CT scanner using TLDs and air kerma - area product (P{sub ka}) as kerma index. Measurements were performed on two protocols used to obtain the image the maxilla-mandible in equipment Gendex GXCB 500: Protocol [GX1] extended diameter and asymmetric beam (14 cm x 8.5 cm - maxilla / mandible) and protocol [GX2] symmetrical beam (8.5 cm x 8.5 cm - maxillary / mandible). Was used LiF dosimeters (TLD 100) inserted into a female anthropomorphic phantom manufactured by Radiology Support Devices. For all protocols evaluated the value of P{sub ka} using a meter Diamentor E2 and PTW system Radcal Rapidose. The results obtained for Effective Dose / P{sub ka} these measurements were separated by protocol image. Protocol [GX1]: 44.5 μSv/478 mGy cm{sup 2}; protocol [GX2]: 54.8 μSv/507 mGy cm{sup 2}. These values indicate that the relationship between the diameter of the image acquired in the protocol [GX1] and the diameter of the image in the protocol [GX2] is equal to 1.65, the Effective Dose for the first protocol has lower value at 18%. P{sub ka} values reveal very similar results between the two protocols, although, common sense leads to the interpretation that imaging protocols with field of view (Fov) of large diameters imply high values of effective dose when compared to small diameters. However, in this particular case, this is not true due to the asymmetrical beam technology. Conclude that for the cases where the scanner uses asymmetric beam to obtain images with large diameters that cover the entire face there are advantages from the point of view of reducing the exposure of patients with respect to the use of symmetrical beam and / or to Fov images with a smaller diameter. (Author)

  2. Assessment of protocols in cone beam CT with symmetric and asymmetric beam using effective dose and Pka

    International Nuclear Information System (INIS)

    Batista, W. O.; Linhares de O, M. V.; Soares, M. R.; Maia, A. F.; Caldas, L. V. E.

    2014-08-01

    The cone beam CT is an emerging technology in dental radiology with significant differences the point of view of design technology between the various manufacturers on the world market. This study aims to evaluate and compare protocols with similar purposes in a cone beam CT scanner using TLDs and air kerma - area product (P ka ) as kerma index. Measurements were performed on two protocols used to obtain the image the maxilla-mandible in equipment Gendex GXCB 500: Protocol [GX1] extended diameter and asymmetric beam (14 cm x 8.5 cm - maxilla / mandible) and protocol [GX2] symmetrical beam (8.5 cm x 8.5 cm - maxillary / mandible). Was used LiF dosimeters (TLD 100) inserted into a female anthropomorphic phantom manufactured by Radiology Support Devices. For all protocols evaluated the value of P ka using a meter Diamentor E2 and PTW system Radcal Rapidose. The results obtained for Effective Dose / P ka these measurements were separated by protocol image. Protocol [GX1]: 44.5 μSv/478 mGy cm 2 ; protocol [GX2]: 54.8 μSv/507 mGy cm 2 . These values indicate that the relationship between the diameter of the image acquired in the protocol [GX1] and the diameter of the image in the protocol [GX2] is equal to 1.65, the Effective Dose for the first protocol has lower value at 18%. P ka values reveal very similar results between the two protocols, although, common sense leads to the interpretation that imaging protocols with field of view (Fov) of large diameters imply high values of effective dose when compared to small diameters. However, in this particular case, this is not true due to the asymmetrical beam technology. Conclude that for the cases where the scanner uses asymmetric beam to obtain images with large diameters that cover the entire face there are advantages from the point of view of reducing the exposure of patients with respect to the use of symmetrical beam and / or to Fov images with a smaller diameter. (Author)

  3. Automatic segmentation of maxillofacial cysts in cone beam CT images.

    Science.gov (United States)

    Abdolali, Fatemeh; Zoroofi, Reza Aghaeizadeh; Otake, Yoshito; Sato, Yoshinobu

    2016-05-01

    Accurate segmentation of cysts and tumors is an essential step for diagnosis, monitoring and planning therapeutic intervention. This task is usually done manually, however manual identification and segmentation is tedious. In this paper, an automatic method based on asymmetry analysis is proposed which is general enough to segment various types of jaw cysts. The key observation underlying this approach is that normal head and face structure is roughly symmetric with respect to midsagittal plane: the left part and the right part can be divided equally by an axis of symmetry. Cysts and tumors typically disturb this symmetry. The proposed approach consists of three main steps as follows: At first, diffusion filtering is used for preprocessing and symmetric axis is detected. Then, each image is divided into two parts. In the second stage, free form deformation (FFD) is used to correct slight displacement of corresponding pixels of the left part and a reflected copy of the right part. In the final stage, intensity differences are analyzed and a number of constraints are enforced to remove false positive regions. The proposed method has been validated on 97 Cone Beam Computed Tomography (CBCT) sets containing various jaw cysts which were collected from various image acquisition centers. Validation is performed using three similarity indicators (Jaccard index, Dice's coefficient and Hausdorff distance). The mean Dice's coefficient of 0.83, 0.87 and 0.80 is achieved for Radicular, Dentigerous and KCOT classes, respectively. For most of the experiments done, we achieved high true positive (TP). This means that a large number of cyst pixels are correctly classified. Quantitative results of automatic segmentation show that the proposed method is more effective than one of the recent methods in the literature. Copyright © 2016 Elsevier Ltd. All rights reserved.

  4. Measurement of cone beam CT coincidence with megavoltage isocentre and image sharpness using the QUASAR(TM) Penta-Guide phantom

    International Nuclear Information System (INIS)

    Sykes, J R; Lindsay, R; Dean, C J; Thwaites, D I; Brettle, D S; Magee, D R

    2008-01-01

    For image-guided radiotherapy (IGRT) systems based on cone beam CT (CBCT) integrated into a linear accelerator, the reproducible alignment of imager to x-ray source is critical to the registration of both the x-ray-volumetric image with the megavoltage (MV) beam isocentre and image sharpness. An enhanced method of determining the CBCT to MV isocentre alignment using the QUASAR(TM) Penta-Guide phantom was developed which improved both precision and accuracy. This was benchmarked against our existing method which used software and a ball-bearing (BB) phantom provided by Elekta. Additionally, a method of measuring an image sharpness metric (MTF 50 ) from the edge response function of a spherical air cavity within the Penta-Guide phantom was developed and its sensitivity was tested by simulating misalignments of the kV imager. Reproducibility testing of the enhanced Penta-Guide method demonstrated a systematic error of 50 for five measurements was 0.278 ± 0.004 lp mm -1 with no applied misalignment. Simulated misalignments exhibited a clear peak in the MTF 50 enabling misalignments greater than 0.4 mm to be detected. The Penta-Guide phantom can be used to precisely measure CBCT-MV coincidence and image sharpness on CBCT-IGRT systems

  5. Dacryocystography using cone beam CT in patients with lacrimal drainage system obstruction.

    OpenAIRE

    Tschopp, Markus; Bornstein, Michael M.; Sendi, Pedram; Jacobs, Reinhilde; Goldblum, David

    2014-01-01

    PURPOSE To assess the usefulness of cone beam CT (CBCT) for dacryocystography (DCG) using either direct syringing or passive application of contrast medium. METHODS Ten consecutive patients with epiphora who had CBCT-DCG in a sitting position were retrospectively analyzed. CBCT-DCGs were performed using 2 techniques: direct syringing with contrast medium or using the passive technique, where patients received 3 drops of contrast medium into the conjunctival sac before CBCT-DCG. Cl...

  6. Geometric Parameters Estimation and Calibration in Cone-Beam Micro-CT

    Directory of Open Access Journals (Sweden)

    Jintao Zhao

    2015-09-01

    Full Text Available The quality of Computed Tomography (CT images crucially depends on the precise knowledge of the scanner geometry. Therefore, it is necessary to estimate and calibrate the misalignments before image acquisition. In this paper, a Two-Piece-Ball (TPB phantom is used to estimate a set of parameters that describe the geometry of a cone-beam CT system. Only multiple projections of the TPB phantom at one position are required, which can avoid the rotation errors when acquiring multi-angle projections. Also, a corresponding algorithm is derived. The performance of the method is evaluated through simulation and experimental data. The results demonstrated that the proposed method is valid and easy to implement. Furthermore, the experimental results from the Micro-CT system demonstrate the ability to reduce artifacts and improve image quality through geometric parameter calibration.

  7. Actively triggered 4d cone-beam CT acquisition.

    Science.gov (United States)

    Fast, Martin F; Wisotzky, Eric; Oelfke, Uwe; Nill, Simeon

    2013-09-01

    4d cone-beam computed tomography (CBCT) scans are usually reconstructed by extracting the motion information from the 2d projections or an external surrogate signal, and binning the individual projections into multiple respiratory phases. In this "after-the-fact" binning approach, however, projections are unevenly distributed over respiratory phases resulting in inefficient utilization of imaging dose. To avoid excess dose in certain respiratory phases, and poor image quality due to a lack of projections in others, the authors have developed a novel 4d CBCT acquisition framework which actively triggers 2d projections based on the forward-predicted position of the tumor. The forward-prediction of the tumor position was independently established using either (i) an electromagnetic (EM) tracking system based on implanted EM-transponders which act as a surrogate for the tumor position, or (ii) an external motion sensor measuring the chest-wall displacement and correlating this external motion to the phase-shifted diaphragm motion derived from the acquired images. In order to avoid EM-induced artifacts in the imaging detector, the authors devised a simple but effective "Faraday" shielding cage. The authors demonstrated the feasibility of their acquisition strategy by scanning an anthropomorphic lung phantom moving on 1d or 2d sinusoidal trajectories. With both tumor position devices, the authors were able to acquire 4d CBCTs free of motion blurring. For scans based on the EM tracking system, reconstruction artifacts stemming from the presence of the EM-array and the EM-transponders were greatly reduced using newly developed correction algorithms. By tuning the imaging frequency independently for each respiratory phase prior to acquisition, it was possible to harmonize the number of projections over respiratory phases. Depending on the breathing period (3.5 or 5 s) and the gantry rotation time (4 or 5 min), between ∼90 and 145 projections were acquired per respiratory

  8. The completeness condition and source orbits for exact image reconstruction in 3D cone-beam CT

    International Nuclear Information System (INIS)

    Mao Xiping; Kang Kejun

    1997-01-01

    The completeness condition for exact image reconstruction in 3D cone-beam CT are carefully analyzed in theory, and discussions about some source orbits which fulfill the completeness condition are followed

  9. Truncation artifact suppression in cone-beam radionuclide transmission CT using maximum likelihood techniques: evaluation with human subjects

    International Nuclear Information System (INIS)

    Manglos, S.H.

    1992-01-01

    Transverse image truncation can be a serious problem for human imaging using cone-beam transmission CT (CB-CT) implemented on a conventional rotating gamma camera. This paper presents a reconstruction method to reduce or eliminate the artifacts resulting from the truncation. The method uses a previously published transmission maximum likelihood EM algorithm, adapted to the cone-beam geometry. The reconstruction method is evaluated qualitatively using three human subjects of various dimensions and various degrees of truncation. (author)

  10. Investigation of the HU-density conversion method and comparison of dose distribution for dose calculation on MV cone beam CT images

    International Nuclear Information System (INIS)

    Kim, Min Joo; Lee, Seu Ran; Suh, Tae Suk

    2011-01-01

    Modern radiation therapy techniques, such as Image-guided radiation therapy (IGRT), Adaptive radiation therapy (ART) has become a routine clinical practice on linear accelerators for the increase the tumor dose conformity and improvement of normal tissue sparing at the same time. For these highly developed techniques, megavoltage cone beam computed tomography (MVCBCT) system produce volumetric images at just one rotation of the x-ray beam source and detector on the bottom of conventional linear accelerator for real-time application of patient condition into treatment planning. MV CBCT image scan be directly registered to a reference CT data set which is usually kilo-voltage fan-beam computed tomography (kVFBCT) on treatment planning system and the registered image scan be used to adjust patient set-up error. However, to use MV CBCT images in radiotherapy, reliable electron density (ED) distribution are required. Patients scattering, beam hardening and softening effect caused by different energy application between kVCT, MV CBCT can cause cupping artifacts in MV CBCT images and distortion of Houns field Unit (HU) to ED conversion. The goal of this study, for reliable application of MV CBCT images into dose calculation, MV CBCT images was modified to correct distortion of HU to ED using the relationship of HU and ED from kV FBCT and MV CBCT images. The HU-density conversion was performed on MV CBCT image set using Dose difference map was showing in Figure 1. Finally, percentage differences above 3% were reduced depending on applying density calibration method. As a result, total error co uld be reduced to under 3%. The present study demonstrates that dose calculation accuracy using MV CBCT image set can be improved my applying HU-density conversion method. The dose calculation and comparison of dose distribution from MV CBCT image set with/without HU-density conversion method was performed. An advantage of this study compared to other approaches is that HU

  11. Investigation of the HU-density conversion method and comparison of dose distribution for dose calculation on MV cone beam CT images

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Min Joo; Lee, Seu Ran; Suh, Tae Suk [Dept. of Biomedical Engineering, The Catholic University of Korea, Bucheon (Korea, Republic of)

    2011-11-15

    Modern radiation therapy techniques, such as Image-guided radiation therapy (IGRT), Adaptive radiation therapy (ART) has become a routine clinical practice on linear accelerators for the increase the tumor dose conformity and improvement of normal tissue sparing at the same time. For these highly developed techniques, megavoltage cone beam computed tomography (MVCBCT) system produce volumetric images at just one rotation of the x-ray beam source and detector on the bottom of conventional linear accelerator for real-time application of patient condition into treatment planning. MV CBCT image scan be directly registered to a reference CT data set which is usually kilo-voltage fan-beam computed tomography (kVFBCT) on treatment planning system and the registered image scan be used to adjust patient set-up error. However, to use MV CBCT images in radiotherapy, reliable electron density (ED) distribution are required. Patients scattering, beam hardening and softening effect caused by different energy application between kVCT, MV CBCT can cause cupping artifacts in MV CBCT images and distortion of Houns field Unit (HU) to ED conversion. The goal of this study, for reliable application of MV CBCT images into dose calculation, MV CBCT images was modified to correct distortion of HU to ED using the relationship of HU and ED from kV FBCT and MV CBCT images. The HU-density conversion was performed on MV CBCT image set using Dose difference map was showing in Figure 1. Finally, percentage differences above 3% were reduced depending on applying density calibration method. As a result, total error co uld be reduced to under 3%. The present study demonstrates that dose calculation accuracy using MV CBCT image set can be improved my applying HU-density conversion method. The dose calculation and comparison of dose distribution from MV CBCT image set with/without HU-density conversion method was performed. An advantage of this study compared to other approaches is that HU

  12. Cone-beam CT hepatic arteriography in chemoembolization for hepatocellular carcinoma: angiographic image quality and its determining factors.

    Science.gov (United States)

    Lee, In Joon; Chung, Jin Wook; Yin, Yong Hu; Kim, Hyo-Cheol; Kim, Young Il; Jae, Hwan Jun; Park, Jae Hyung

    2014-09-01

    To analyze image quality and the factors that determine it for cone-beam computed tomography (CT) hepatic arteriography in chemoembolization for hepatocellular carcinoma (HCC). From September 2009-December 2010, 399 consecutive patients referred for chemoembolization of HCC were scheduled for cone-beam CT scan. There were 12 patients (3%) excluded because of difficulty with breath-hold. Of the 387 patients who underwent cone-beam CT hepatic arteriography, 100 patients were ultimately included in the study according to inclusion criteria. Maximum intensity projection images were scored for image quality of each segmental hepatic artery. Potential determining factors for image quality were diaphragmatic motion, portal vein enhancement, and hepatic artery-to-parenchyma enhancement ratio. The flow rate of contrast media, x-ray delay, and location of the catheter tip were also evaluated. It was possible to trace at least subsegmental hepatic arteries in 625 of 700 segments (89.3%) on cone-beam CT hepatic arteriography. Diaphragmatic motion, prominent portal vein enhancement, and low hepatic artery-to-parenchyma enhancement ratio worsened image quality (P cone-beam CT hepatic arteriography images was good enough to trace subsegmental hepatic arteries at a minimum. Respiratory and cardiac motion, portal vein enhancement, and hepatic artery-to-parenchyma enhancement ratio significantly affected the image quality of cone-beam CT hepatic arteriography. Copyright © 2014 SIR. Published by Elsevier Inc. All rights reserved.

  13. Bifid mandibular canal: confirmation of limited cone beam CT findings by gross anatomical and histological investigations

    Science.gov (United States)

    Fukami, K; Shiozaki, K; Mishima, A; Kuribayashi, A; Hamada, Y; Kobayashi, K

    2012-01-01

    Objectives The aims of this study were (1) to assess the validity of limited cone beam CT (CBCT) in detecting the distribution of bifid mandibular canals in the retromolar region by comparing its findings with those of panoramic radiography and spiral CT imaging, and (2) to confirm the contents of such canals depicted on limited CBCT images by using gross anatomical and histological methods. Methods Bilateral bifid mandibular canals of a Japanese cadaver were investigated. The canals depicted on panoramic radiography, spiral CT and limited CBCT images were compared. Cross-sectional limited CBCT images of these canals were compared with gross anatomical sections of the mandible and their contents were confirmed histologically. Results The spiral CT and limited CBCT images showed the bilateral bifid mandibular canals in the retromolar region whereas the panoramic radiographs indicated the presence of only the left bifid mandibular canal. The canal distribution was more distinct in the limited CBCT images than in the spiral CT images and the cross-sectional limited CBCT images were consistent with the gross anatomical sections. Histologically, the canals contained several nerve bundles and arteries among which the largest nerve and artery were of a similar size. Conclusion Limited CBCT is valuable for assessing the distribution of bifid mandibular canals. It is clinically significant to accurately localize a bifid mandibular canal of the retromolar region because it contains a nerve bundle and artery. PMID:22116121

  14. Comparison of effective dose for imaging of mandible between multi-detector CT and cone-beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Jeong, Dae Kyo; Lee, Sang Chul; Huh, Kyung Hoe; Yi, Won Jin; Lee, Sam Sun; Choi, Soon Chul [School of Dentistry, Seoul National University, Seoul (Korea, Republic of)

    2012-06-15

    The aim of this study was to compare the effective dose for imaging of mandible between multi-detector computed tomography (MDCT) and cone-beam computed tomography (CBCT). An MDCT with low dose technique was also compared with them. Thermoluminescent dosimeter (TLD) chips were placed at 25 organ sites of an anthropomorphic phantom. The mandible of the phantom was exposed using 2 different types of MDCT units (Somatom Sensation 10 for standard-dose MDCT, Somatom Emotion 6 for low-dose MDCT) and 3 different CBCT units (AZ3000CT, Implagraphy, and Kavo 3D eXaM). The radiation absorbed dose was measured and the effective dose was calculated according to the ICRP 2007 report. The effective dose was the highest for Somatom Sensation 10 (425.84 {mu}Sv), followed by AZ3000CT (332.4 {mu}Sv), Somatom Emotion 6 (199.38 {mu}Sv), and 3D eXaM (111.6 {mu}Sv); it was the lowest for Implagraphy (83.09 {mu}Sv). The CBCT showed significant variation in dose level with different device. The effective doses of MDCTs were not significantly different from those of CBCTs for imaging of mandible. The effective dose of MDCT could be markedly decreased by using the low-dose technique.

  15. GPU-accelerated regularized iterative reconstruction for few-view cone beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Matenine, Dmitri, E-mail: dmitri.matenine.1@ulaval.ca [Département de physique, de génie physique et d’optique, Université Laval, Québec, Québec G1V 0A6 (Canada); Goussard, Yves, E-mail: yves.goussard@polymtl.ca [Département de génie électrique/Institut de génie biomédical, École Polytechnique de Montréal, C.P. 6079, succ. Centre-ville, Montréal, Québec H3C 3A7 (Canada); Després, Philippe, E-mail: philippe.despres@phy.ulaval.ca [Département de physique, de génie physique et d’optique and Centre de recherche sur le cancer, Université Laval, Québec, Québec G1V 0A6 (Canada)

    2015-04-15

    Purpose: The present work proposes an iterative reconstruction technique designed for x-ray transmission computed tomography (CT). The main objective is to provide a model-based solution to the cone-beam CT reconstruction problem, yielding accurate low-dose images via few-views acquisitions in clinically acceptable time frames. Methods: The proposed technique combines a modified ordered subsets convex (OSC) algorithm and the total variation minimization (TV) regularization technique and is called OSC-TV. The number of subsets of each OSC iteration follows a reduction pattern in order to ensure the best performance of the regularization method. Considering the high computational cost of the algorithm, it is implemented on a graphics processing unit, using parallelization to accelerate computations. Results: The reconstructions were performed on computer-simulated as well as human pelvic cone-beam CT projection data and image quality was assessed. In terms of convergence and image quality, OSC-TV performs well in reconstruction of low-dose cone-beam CT data obtained via a few-view acquisition protocol. It compares favorably to the few-view TV-regularized projections onto convex sets (POCS-TV) algorithm. It also appears to be a viable alternative to full-dataset filtered backprojection. Execution times are of 1–2 min and are compatible with the typical clinical workflow for nonreal-time applications. Conclusions: Considering the image quality and execution times, this method may be useful for reconstruction of low-dose clinical acquisitions. It may be of particular benefit to patients who undergo multiple acquisitions by reducing the overall imaging radiation dose and associated risks.

  16. Non-rigid CT/CBCT to CBCT registration for online external beam radiotherapy guidance

    Science.gov (United States)

    Zachiu, Cornel; de Senneville, Baudouin Denis; Tijssen, Rob H. N.; Kotte, Alexis N. T. J.; Houweling, Antonetta C.; Kerkmeijer, Linda G. W.; Lagendijk, Jan J. W.; Moonen, Chrit T. W.; Ries, Mario

    2018-01-01

    Image-guided external beam radiotherapy (EBRT) allows radiation dose deposition with a high degree of accuracy and precision. Guidance is usually achieved by estimating the displacements, via image registration, between cone beam computed tomography (CBCT) and computed tomography (CT) images acquired at different stages of the therapy. The resulting displacements are then used to reposition the patient such that the location of the tumor at the time of treatment matches its position during planning. Moreover, ongoing research aims to use CBCT-CT image registration for online plan adaptation. However, CBCT images are usually acquired using a small number of x-ray projections and/or low beam intensities. This often leads to the images being subject to low contrast, low signal-to-noise ratio and artifacts, which ends-up hampering the image registration process. Previous studies addressed this by integrating additional image processing steps into the registration procedure. However, these steps are usually designed for particular image acquisition schemes, therefore limiting their use on a case-by-case basis. In the current study we address CT to CBCT and CBCT to CBCT registration by the means of the recently proposed EVolution registration algorithm. Contrary to previous approaches, EVolution does not require the integration of additional image processing steps in the registration scheme. Moreover, the algorithm requires a low number of input parameters, is easily parallelizable and provides an elastic deformation on a point-by-point basis. Results have shown that relative to a pure CT-based registration, the intrinsic artifacts present in typical CBCT images only have a sub-millimeter impact on the accuracy and precision of the estimated deformation. In addition, the algorithm has low computational requirements, which are compatible with online image-based guidance of EBRT treatments.

  17. Noise power properties of a cone-beam CT system for breast cancer detection.

    Science.gov (United States)

    Yang, Kai; Kwan, Alexander L C; Huang, Shih-Ying; Packard, Nathan J; Boone, John M

    2008-12-01

    The noise power properties of a cone-beam computed tomography (CT) system dedicated for breast cancer detection were investigated. Uniform polyethylene cylinders of various diameters were scanned under different system acquisition conditions. Noise power spectra were calculated from difference data generated by subtraction between two identical scans. Multidimensional noise power spectra (NPS) were used as the metric to evaluate the noise properties of the breast CT (bCT) under different system acquisition and reconstruction conditions. A comprehensive investigation of the noise properties was performed in regard to system acquisition parameters including kVp, mA, number of cone-beam projection images used, cone angle, and object size. The influence on reconstruction parameters including interpolation method, reconstruction filter, field of view, matrix size, and slice thickness were also studied. Under certain conditions, the zero-dimensional NPS (image variance) was used as a quantitative index to compare the influence from different scan parameters, especially the radiation dose. If the total scan dose is changed by linearly changing the total number of projection images while the dose per frame is kept constant, the noise power has a linear relationship with the reciprocal of the total dose. If the total scan dose is changed by linearly changing the dose per frame while the total number of projection images is kept constant, the noise power has a quadratic relationship with the reciprocal of the total dose. With the same amount of total dose, using fewer projection images results in lower image noise power in the CT image. Quantitative results from this noise power analysis provide guidance for the bCT system operation, optimization, and data reconstruction.

  18. Interleaved acquisition for cross scatter avoidance in dual cone-beam CT.

    Science.gov (United States)

    Giles, William; Bowsher, James; Li, Hao; Yin, Fang-Fang

    2012-12-01

    Cone-beam x-ray imaging with flat panel detectors is used for target localization in image guided radiation therapy. This imaging includes cone-beam computed tomography (CBCT) and planar imaging. Use of two orthogonal x-ray systems could reduce imaging time for CBCT, provide simultaneous orthogonal views in planar imaging, facilitate dual-energy methods, and be useful in alleviating cone-beam artifacts by providing two axially offset focal-spot trajectories. However, the potential advantages of a second cone-beam system come at the cost of cross scatter, i.e., scatter of photons originating from one tube into the noncorresponding detector. Herein, cross scatter is characterized for dual cone-beam imaging, and a method for avoiding cross scatter is proposed and evaluated. A prototype dual-source CBCT system has been developed that models the geometry of a gantry-mounted kV imaging device used in radiation therapy. Cross scatter was characterized from 70 to 145 kVp in projections and reconstructed images using this system and three cylindrical phantoms (15, 20, and 30 cm) with a common Catphan core. A novel strategy for avoiding cross scatter in dual CBCT was developed that utilized interleaved data acquisition on each imaging chain. Interleaving, while maintaining similar angular sampling, can be achieved by either doubling the data acquisition rate or, as presented herein, halving the rotation speed. The ratio of cross scatter to the total detected signal was found to be as high as 0.59 in a 30 cm diameter phantom. The measured scatter-to-primary ratio in some cases exceeded 4. In the 30 cm phantom, reconstructed contrast was reduced across all ROIs by an average of 48.7% when cross scatter was present. These cross-scatter degradations were almost entirely avoided by the method of interleaved exposures. Cross scatter is substantial in dual cone-beam imaging, but its effects can be largely removed by interleaved acquisition, which can be achieved at the same angular

  19. Extended localization and adaptive dose calculation using HU corrected cone beam CT: Phantom study.

    Science.gov (United States)

    Rafic, K Mohamathu; Amalan, S; Timothy Peace, B S; Ravindran, B Paul

    2018-01-01

    The practicability of computing dose calculation on cone beam CT (CBCT) has been widely investigated. In most clinical scenarios, the craniocaudal scanning length of CBCT is found to be inadequate for localization. This study aims to explore extended tomographic localization and adaptive dose calculation strategies using Hounsfield unit (HU) corrected CBCT image sets. Planning CT (pCT) images of the Rando phantom (T 12 -to-midthigh) were acquired with pelvic-protocol using Biograph CT-scanner. Similarly, half-fan CBCT were acquired with fixed parameters using Clinac2100C/D linear accelerator integrated with an on-board imager with 2-longitudinal positions of the table. For extended localization and dose calculation, two stitching strategies viz., one with "penumbral-overlap" (S 1 ) and the other with "no-overlap" (S 2 ) and a local HU-correction technique were performed using custom-developed MATLAB scripts. Fluence modulated treatment plans computed on pCT were mapped with stitched CBCT and the dosimetric analyses such as dose-profile comparison, 3D-gamma (γ) evaluation and dose-volume histogram (DVH) comparison were performed. Localizing scanning length of CBCT was extended by up to 15 cm and 16 cm in S 1 and S 2 strategies, respectively. Treatment plan mapping resulted in minor variations in the volumes of delineated structures and the beam centre co-ordinates. While the former showed maximum variations of -1.4% and -1.6%, the latter showed maximum of 1.4 mm and 2.7 mm differences in anteroposterior direction in S 1 and S 2 protocols, respectively. Dosimetric evaluations viz., dose profile and DVH comparisons were found to be in agreement with one another. In addition, γ-evaluation results showed superior pass-rates (≥98.5%) for both 3%/3 mm dose-difference (DD) and distance-to-agreement (DTA) and 2%/2 mm DD/DTA criteria with desirable dosimetric accuracy. Cone beam tomographic stitching and local HU-correction strategies developed to facilitate

  20. Cone-beam CT angiography of the thorax. An experimental study

    Energy Technology Data Exchange (ETDEWEB)

    Yoshida, Katsuya; Shimada, Kazuhiro [Chiba Univ. (Japan). School of Medicine; Tadokoro, Hiroyuki (and others)

    1999-10-01

    The authors recently developed a cone-beam computed tomography (CT) scanner and this report presents their evaluation of its potential for thoracic vascular imaging. An X-ray tube and a video-fluoroscopic system were rotated around the objects and 360 projected images were collected in a 12-s scan. Each image was digitized and a 3 dimensional (D) image (256 x 256 x 256 voxel volume with a voxel dimension of 0.9 x 0.9 x 0.9 mm) was reconstructed. Two different 3D-CT angiographies were investigated in 2 pigs: right atriography and thoracic aortography. Each pig was anesthetized, mechanically ventilated and positioned within the scanner. Contrast agent was infused through the right atrium or the aortic root at a rate of 3 ml/s during the scan. The right atriography scan clearly delineated the anatomy of the pulmonary artery, heart chambers and thoracic aorta. The thoracic aortography scan also clearly delineated the aortic anatomy including the internal thoracic and intercostal arteries. In conclusion, cone-beam CT angiography is potentially useful for thoracic vascular imaging. (author)

  1. Cone-beam CT angiography of the thorax. An experimental study

    International Nuclear Information System (INIS)

    Yoshida, Katsuya; Shimada, Kazuhiro; Tadokoro, Hiroyuki

    1999-01-01

    The authors recently developed a cone-beam computed tomography (CT) scanner and this report presents their evaluation of its potential for thoracic vascular imaging. An X-ray tube and a video-fluoroscopic system were rotated around the objects and 360 projected images were collected in a 12-s scan. Each image was digitized and a 3 dimensional (D) image (256 x 256 x 256 voxel volume with a voxel dimension of 0.9 x 0.9 x 0.9 mm) was reconstructed. Two different 3D-CT angiographies were investigated in 2 pigs: right atriography and thoracic aortography. Each pig was anesthetized, mechanically ventilated and positioned within the scanner. Contrast agent was infused through the right atrium or the aortic root at a rate of 3 ml/s during the scan. The right atriography scan clearly delineated the anatomy of the pulmonary artery, heart chambers and thoracic aorta. The thoracic aortography scan also clearly delineated the aortic anatomy including the internal thoracic and intercostal arteries. In conclusion, cone-beam CT angiography is potentially useful for thoracic vascular imaging. (author)

  2. Roman Pot Insertions in High-Intensity Beams for the CT-PPS Project at LHC

    CERN Document Server

    Deile, Mario; Mereghetti, Alessio; Mirarchi, Daniele; Redaelli, Stefano; Salvachua, Belen; Salvant, Benoit; Valentino, Gianluca

    2016-01-01

    The CMS-TOTEM Precision Proton Spectrometer (CT-PPS) at the LHC IP5 aims at exploring diffractive physics at high luminosity in standard LHC fills. It is based on 14 Roman Pots (RPs), designed to host tracking and time-of-flight detectors for measuring the kinematics of leading protons. To reach the physics goals, the RPs will finally have to approach the beams to distances of 15 beam σs (i.e. ~1.5 mm) or closer. After problems with showers and impedance heating in first high-luminosity RP insertions in 2012, the LS1 of LHC was used for upgrades in view of impedance minimisation and for adding new collimators to intercept RP-induced showers. In 2015 the effectiveness of these improvements was shown by successfully inserting the RPs in all LHC beam intensity steps to a first-phase distance of ~25 σs. This contribution reviews the measurements of debris showers and impedance effects, i.e. the data from Beam Loss Monitors, beam vacuum gauges and temperature sensors. The dependences of the observables on the lu...

  3. On board imaging with cone beam C.B.C.T. kV VARIAN{sup TM}: Montauban's radiation therapy department experience; Systeme d'imagerie par tomographie conique de basse energie (kV)de Varian{sup TM}: experience de Montauban

    Energy Technology Data Exchange (ETDEWEB)

    Dudouet, P.; Boutry, C.; Mounie, G.; Thouveny, F.; Redon, A. [Clinique du Pont-de-Chaume, Groupe Oncorad Garonne, Service d' Oncologie, 82 - Montauban (France); Latorzeff, I. [Clinique Pasteur, L' Atrium, Service de Radiotherapie, Groupe Oncorad Garonne, 31 - Toulouse (France)

    2009-09-15

    Purpose: To describe our practice day to day with a Variant linac 'Clinac 2100' fully equipped with an On Board Imager (O.B.I.) for patients with prostate cancer. Materials and methods: A volumetric and dosimetric study was performed in 2006 using Artiview software (Aquilab) for nine patients and 76 Cone Beam CT kV (C.B.C.T. kV). We have contoured targets and organs at risk from C.B.C.T. kV slides acquisitions. Second, we achieved a dose-volume histogram (D.V.H.) study for a patient treated in 2007 with I.M.R.T. technique in comparison with the 2006 study. Results: 2006 analysis: The study showed a very important variability of organ measurements. Seminal vesicles were strongly influenced by adjacent organs; observed differences for prostate could be explained by contouring uncertainty on the apex. Inter-sessions motions could be observed for bladder, rectum and seminal vesicles (S.V.). Part of prostate volume not encompassed by P.T.V. is about 2.5%; V.S. volume outside P.T.V. is about 35%. Tumoral conformation index (T.C.I.) is inferior to 97.5% in 22% of all cases. Anteroposterior displacements of the prostate barycentre is superior to 5 mm. From this analysis, we recommended the strict respect of hygieno-dietetic rules, and we have adapted the system settings for better immobilization, which were applied for the 2007 study. For the 2007 analysis, since April 2007, most of patients are treated with I.M.R.T. for prostate cancer, at the second part of the radiation therapy to encompass only the prostate volume. Dose-volume histograms showed a great spreading out for 2006 patients, and not for the 2007 patient. Conclusions: Intensity modulation radiotherapy (I.M.R.T.) and image guided radiotherapy (I.G.R.T.) should permit a margin reduction for P.T.V.. Strict respect of hygieno-dietetics rules is necessary to avoid rectal distension and local recurrence. (authors)

  4. A combination-weighted Feldkamp-based reconstruction algorithm for cone-beam CT

    International Nuclear Information System (INIS)

    Mori, Shinichiro; Endo, Masahiro; Komatsu, Shuhei; Kandatsu, Susumu; Yashiro, Tomoyasu; Baba, Masayuki

    2006-01-01

    The combination-weighted Feldkamp algorithm (CW-FDK) was developed and tested in a phantom in order to reduce cone-beam artefacts and enhance cranio-caudal reconstruction coverage in an attempt to improve image quality when utilizing cone-beam computed tomography (CBCT). Using a 256-slice cone-beam CT (256CBCT), image quality (CT-number uniformity and geometrical accuracy) was quantitatively evaluated in phantom and clinical studies, and the results were compared to those obtained with the original Feldkamp algorithm. A clinical study was done in lung cancer patients under breath holding and free breathing. Image quality for the original Feldkamp algorithm is degraded at the edge of the scan region due to the missing volume, commensurate with the cranio-caudal distance between the reconstruction and central planes. The CW-FDK extended the reconstruction coverage to equal the scan coverage and improved reconstruction accuracy, unaffected by the cranio-caudal distance. The extended reconstruction coverage with good image quality provided by the CW-FDK will be clinically investigated for improving diagnostic and radiotherapy applications. In addition, this algorithm can also be adapted for use in relatively wide cone-angle CBCT such as with a flat-panel detector CBCT

  5. Passive breath gating equipment for cone beam CT-guided RapidArc gastric cancer treatments.

    Science.gov (United States)

    Hu, Weigang; Li, Guichao; Ye, Jinsong; Wang, Jiazhou; Peng, Jiayuan; Gong, Min; Yu, Xiaoli; Studentski, Matthew T; Xiao, Ying; Zhang, Zhen

    2015-01-01

    To report preliminary results of passive breath gating (PBG) equipment for cone-beam CT image-guided gated RapidArc gastric cancer treatments. Home-developed PBG equipment integrated with the real-time position management system (RPM) for passive patient breath hold was used in CT simulation, online partial breath hold (PBH) CBCT acquisition, and breath-hold gating (BHG) RapidArc delivery. The treatment was discontinuously delivered with beam on during BH and beam off for free breathing (FB). Pretreatment verification PBH CBCT was obtained with the PBG-RPM system. Additionally, the reproducibility of the gating accuracy was evaluated. A total of 375 fractions of breath-hold gating RapidArc treatments were successfully delivered and 233 PBH CBCTs were available for analysis. The PBH CBCT images were acquired with 2-3 breath holds and 1-2 FB breaks. The imaging time was the same for PBH CBCT and conventional FB CBCT (60s). Compared to FB CBCT, the motion artifacts seen in PBH CBCT images were remarkably reduced. The average BHG RapidArc delivery time was 103 s for one 270-degree arc and 269 s for two full arcs. The PBG-RPM based PBH CBCT verification and BHG RapidArc delivery was successfully implemented clinically. The BHG RapidArc treatment was accomplished using a conventional RapidArc machine with high delivery efficiency. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  6. GPU-Based 3D Cone-Beam CT Image Reconstruction for Large Data Volume

    Directory of Open Access Journals (Sweden)

    Xing Zhao

    2009-01-01

    Full Text Available Currently, 3D cone-beam CT image reconstruction speed is still a severe limitation for clinical application. The computational power of modern graphics processing units (GPUs has been harnessed to provide impressive acceleration of 3D volume image reconstruction. For extra large data volume exceeding the physical graphic memory of GPU, a straightforward compromise is to divide data volume into blocks. Different from the conventional Octree partition method, a new partition scheme is proposed in this paper. This method divides both projection data and reconstructed image volume into subsets according to geometric symmetries in circular cone-beam projection layout, and a fast reconstruction for large data volume can be implemented by packing the subsets of projection data into the RGBA channels of GPU, performing the reconstruction chunk by chunk and combining the individual results in the end. The method is evaluated by reconstructing 3D images from computer-simulation data and real micro-CT data. Our results indicate that the GPU implementation can maintain original precision and speed up the reconstruction process by 110–120 times for circular cone-beam scan, as compared to traditional CPU implementation.

  7. Clinical application of 3D reconstruction of tracheobronchial tree with electron beam CT

    International Nuclear Information System (INIS)

    Yao Zhenwei; Shen Tianzhen

    2002-01-01

    Objective: To explore the clinical promise of CT 3D reconstruction of tracheobronchial tree (TBT) by analyzing 73 cases retrospectively. Methods: All the 73 cases were collected from October 1997 to February 2000, who were scanned by EBCT with 130 kV and 630 mA. The scanning method was continuous volume scan, the slice thickness were 3 mm or 1.5 mm. All cross-sectional images were transmitted to the INSIGHT workstation and reconstructed with SSD (shaded surface display), and the threshold setting were -500 to -300 HU. Results: 3D reconstruction of TBT with EBCT could reveal the abnormal changes of TBT by many kinds of diseases including central cancer, inflammation, bronchiectasis, saber-sheath trachea, trachea cancer, congenital disorders, post-surgical changes of lung cancer, and stenoses by adjacent benign or malignant diseases. It could be used to locate the stenoses and measure stenotic extent. Of the 35 central cancer cases with 3D reconstruction, 6 cases were pestle obstructed, 15 cases cone obstructed, 5 cases interrupted irregularly, 8 cases with eccentric stenoses, and 1 case with right stem destroyed and right upper lobe bronchus obstructed. Conclusion: 3D reconstruction of TBT has characteristic sign in the diagnosis or differential diagnosis of central airway's benign or malignant stenoses, and it is of instructional value in clinical use

  8. Cone-beam CT: an additional imaging tool in the interventional treatment and management of low-flow vascular malformations.

    Science.gov (United States)

    Lightfoot, Christopher B; Ju, Yang; Dubois, Josée; Abdolell, Mohamed; Giroux, Marie-France; Gilbert, Patrick; Therasse, Eric; Oliva, Vincent; Soulez, Gilles

    2013-07-01

    To evaluate the impact of cone-beam computed tomography (CT) during sclerotherapy of low-flow vascular malformations. Eighty-seven cone-beam CT examinations were acquired during 81 sclerotherapy treatments of low-flow malformations in 48 patients: 81 were performed to evaluate sclerosing agent diffusion and six were performed to evaluate needle or catheter positioning before injection of therapeutic agent. Image quality was rated by two observers. Clinical impact of cone-beam CT in the assessment of therapeutic agent diffusion, needle or catheter positioning, subsequent treatment planning, and complication detection was evaluated. The κ-statistic was used to assess interobserver reliability and proportions, with associated 95% confidence intervals (CIs). All cone-beam CT images were successfully acquired. Image quality was rated as excellent or good for the majority of studies, with substantial interobserver reliability (κ = 0.648). Cone-beam CT studies improved assessment of therapeutic agent diffusion in 83% of cases (67 of 81; 95% CI, 75%-91%) for observer 1, who had access to ultrasound, fluoroscopic, and digital subtraction angiographic (DSA) imaging, and in 95% of cases (77 of 81; 95% CI, 90%-100%) for observer 2, who had access to only stored fluoroscopic spot radiographs and DSA images. Cone-beam CT impacted planning of the next treatment session in 49% of cases (40 of 81; 95% CI, 38%-60%). In 7% of cases (six of 81; 95% CI, 1%-13%), complications such as migration of therapeutic agent or compression of upper airways were detected that were not seen with other imaging. Cone-beam CT can be a useful adjunctive imaging tool, providing information to help decision-making during percutaneous sclerotherapy and ongoing management of low-flow vascular malformations. Copyright © 2013 SIR. Published by Elsevier Inc. All rights reserved.

  9. Integration of PET-CT and cone-beam CT for image-guided radiotherapy with high image quality and registration accuracy

    Science.gov (United States)

    Wu, T.-H.; Liang, C.-H.; Wu, J.-K.; Lien, C.-Y.; Yang, B.-H.; Huang, Y.-H.; Lee, J. J. S.

    2009-07-01

    Hybrid positron emission tomography-computed tomography (PET-CT) system enhances better differentiation of tissue uptake of 18F-fluorodeoxyglucose (18F-FDG) and provides much more diagnostic value in the non-small-cell lung cancer and nasopharyngeal carcinoma (NPC). In PET-CT, high quality CT images not only offer diagnostic value on anatomic delineation of the tissues but also shorten the acquisition time for attenuation correction (AC) compared with PET-alone imaging. The linear accelerators equipped with the X-ray cone-beam computed tomography (CBCT) imaging system for image-guided radiotherapy (IGRT) provides excellent verification on position setup error. The purposes of our study were to optimize the CT acquisition protocols of PET-CT and to integrate the PET-CT and CBCT for IGRT. The CT imaging parameters were modified in PET-CT for increasing the image quality in order to enhance the diagnostic value on tumour delineation. Reproducibility and registration accuracy via bone co-registration algorithm between the PET-CT and CBCT were evaluated by using a head phantom to simulate a head and neck treatment condition. Dose measurement in computed tomography dose index (CTDI) was also estimated. Optimization of the CT acquisition protocols of PET-CT was feasible in this study. Co-registration accuracy between CBCT and PET-CT on axial and helical modes was in the range of 1.06 to 2.08 and 0.99 to 2.05 mm, respectively. In our result, it revealed that the accuracy of the co-registration with CBCT on helical mode was more accurate than that on axial mode. Radiation doses in CTDI were 4.76 to 18.5 mGy and 4.83 to 18.79 mGy on axial and helical modes, respectively. Registration between PET-CT and CBCT is a state-of-the-art registration technology which could provide much information on diagnosis and accurate tumour contouring on radiotherapy while implementing radiotherapy procedures. This novelty technology of PET-CT and cone-beam CT integration for IGRT may have a

  10. Integration of PET-CT and cone-beam CT for image-guided radiotherapy with high image quality and registration accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Wu, T-H [Department of Medical Imaging and Radiological Sciences, Chung Shan Medical University, No. 110, Sec.1, Jianguo N.Rd, Taichung City 40201, Taiwan (China); Liang, C-H [Agfa Healthcare Systems Taiwan Co., Ltd., 6F, 237 Sung Chiang Road, Taipei, 104 Taiwan (China); Wu, J-K [Division of Radiation Oncology, Department of Oncology, and Cancer Research Center, National Taiwan University Hospital, No.7 Chung San South Road, Taipei, 104 Taiwan (China); Lien, C-Y [Institute of Biomedical Engineering, National Yang Ming University, No. 155, Sec.2, Linong Street, Taipei, 112 Taiwan (China); Yang, B-H; Lee, J J S [Department of Biomedical Imaging and Radiological Sciences, National Yang Ming University, No. 155, Sec.2, Linong Street, Taipei, 112 Taiwan (China); Huang, Y-H [Department of Medical Imaing and Radiological Sciences, I-Shou University, No. 8, Yida Rd., Yanchao Township, Kaohsiung County 82445, Taiwan (China)], E-mail: jslee@ym.edu.tw

    2009-07-15

    Hybrid positron emission tomography-computed tomography (PET-CT) system enhances better differentiation of tissue uptake of {sup 18}F-fluorodeoxyglucose ({sup 18}F-FDG) and provides much more diagnostic value in the non-small-cell lung cancer and nasopharyngeal carcinoma (NPC). In PET-CT, high quality CT images not only offer diagnostic value on anatomic delineation of the tissues but also shorten the acquisition time for attenuation correction (AC) compared with PET-alone imaging. The linear accelerators equipped with the X-ray cone-beam computed tomography (CBCT) imaging system for image-guided radiotherapy (IGRT) provides excellent verification on position setup error. The purposes of our study were to optimize the CT acquisition protocols of PET-CT and to integrate the PET-CT and CBCT for IGRT. The CT imaging parameters were modified in PET-CT for increasing the image quality in order to enhance the diagnostic value on tumour delineation. Reproducibility and registration accuracy via bone co-registration algorithm between the PET-CT and CBCT were evaluated by using a head phantom to simulate a head and neck treatment condition. Dose measurement in computed tomography dose index (CTDI) was also estimated. Optimization of the CT acquisition protocols of PET-CT was feasible in this study. Co-registration accuracy between CBCT and PET-CT on axial and helical modes was in the range of 1.06 to 2.08 and 0.99 to 2.05 mm, respectively. In our result, it revealed that the accuracy of the co-registration with CBCT on helical mode was more accurate than that on axial mode. Radiation doses in CTDI were 4.76 to 18.5 mGy and 4.83 to 18.79 mGy on axial and helical modes, respectively. Registration between PET-CT and CBCT is a state-of-the-art registration technology which could provide much information on diagnosis and accurate tumour contouring on radiotherapy while implementing radiotherapy procedures. This novelty technology of PET-CT and cone-beam CT integration for IGRT

  11. SU-E-I-10: Putting Teeth into Your CT Dosimetry Program: Approaches to Cone- Beam Dental/Maxillofacial CT Dosimetry.

    Science.gov (United States)

    Blackburn, T; Gallet, J; Guild, J; Arbique, G; Anderson, J A

    2012-06-01

    To review and compare different approaches to the problem of dosimetry for limited field-of-view (FOV) cone beam CT devices for dental and maxillofacial applications. The determination of patient doses from specialized, cone-beam CT devices for dental and maxillofacial work requires medical physicists to re-evaluate their dosimetry methods. These devices work in cone-beam geometry, with an axial field dimension on the order of the lengths of the standard head CTDI phantom and pencil ionization chamber. They may also utilize less than 360 degree scans, resulting in asymmetrical radiation distributions. This operating regime is far from that for which conventional CT dosimetry was designed, and alternative approaches must be considered. The alternatives include extensions of conventional CT dosimetry currently used for large axial FOV scanners (e.g. the extended CTDI parameter (CTDIe) for the Toshiba Aquillion One with 160 mm axial FOV) and the new method based on point dosimetry measurements recently formalized in AAPM Report TG-111. Conventional, modified-conventional, and TG-111 dosimetry measurements are used in two CT dose phantoms (adult head and pediatric head) to obtain dose indices for the Planmeca ProMax 3D Max dental CT scanner. Surface dose maps are generated using radiochromic film for correlation with the chamber dosimetry. Results for the three dosimetry approaches are compared for the specific case of the ProMax 3D Max scanner. Strengths and weaknesses of the three measurement paradigms for this type of application are compared. The increasing availability of specialized scanners operating in full cone-beam mode will require the clinical medical physicist to be conversant with extensions to the CT dose index methodology suitable for this equipment. © 2012 American Association of Physicists in Medicine.

  12. SimDoseCT: dose reporting software based on Monte Carlo simulation for a 320 detector-row cone-beam CT scanner and ICRP computational adult phantoms

    Science.gov (United States)

    Cros, Maria; Joemai, Raoul M. S.; Geleijns, Jacob; Molina, Diego; Salvadó, Marçal

    2017-08-01

    This study aims to develop and test software for assessing and reporting doses for standard patients undergoing computed tomography (CT) examinations in a 320 detector-row cone-beam scanner. The software, called SimDoseCT, is based on the Monte Carlo (MC) simulation code, which was developed to calculate organ doses and effective doses in ICRP anthropomorphic adult reference computational phantoms for acquisitions with the Aquilion ONE CT scanner (Toshiba). MC simulation was validated by comparing CTDI measurements within standard CT dose phantoms with results from simulation under the same conditions. SimDoseCT consists of a graphical user interface connected to a MySQL database, which contains the look-up-tables that were generated with MC simulations for volumetric acquisitions at different scan positions along the phantom using any tube voltage, bow tie filter, focal spot and nine different beam widths. Two different methods were developed to estimate organ doses and effective doses from acquisitions using other available beam widths in the scanner. A correction factor was used to estimate doses in helical acquisitions. Hence, the user can select any available protocol in the Aquilion ONE scanner for a standard adult male or female and obtain the dose results through the software interface. Agreement within 9% between CTDI measurements and simulations allowed the validation of the MC program. Additionally, the algorithm for dose reporting in SimDoseCT was validated by comparing dose results from this tool with those obtained from MC simulations for three volumetric acquisitions (head, thorax and abdomen). The comparison was repeated using eight different collimations and also for another collimation in a helical abdomen examination. The results showed differences of 0.1 mSv or less for absolute dose in most organs and also in the effective dose calculation. The software provides a suitable tool for dose assessment in standard adult patients undergoing CT

  13. SimDoseCT: dose reporting software based on Monte Carlo simulation for a 320 detector-row cone-beam CT scanner and ICRP computational adult phantoms.

    Science.gov (United States)

    Cros, Maria; Joemai, Raoul M S; Geleijns, Jacob; Molina, Diego; Salvadó, Marçal

    2017-07-17

    This study aims to develop and test software for assessing and reporting doses for standard patients undergoing computed tomography (CT) examinations in a 320 detector-row cone-beam scanner. The software, called SimDoseCT, is based on the Monte Carlo (MC) simulation code, which was developed to calculate organ doses and effective doses in ICRP anthropomorphic adult reference computational phantoms for acquisitions with the Aquilion ONE CT scanner (Toshiba). MC simulation was validated by comparing CTDI measurements within standard CT dose phantoms with results from simulation under the same conditions. SimDoseCT consists of a graphical user interface connected to a MySQL database, which contains the look-up-tables that were generated with MC simulations for volumetric acquisitions at different scan positions along the phantom using any tube voltage, bow tie filter, focal spot and nine different beam widths. Two different methods were developed to estimate organ doses and effective doses from acquisitions using other available beam widths in the scanner. A correction factor was used to estimate doses in helical acquisitions. Hence, the user can select any available protocol in the Aquilion ONE scanner for a standard adult male or female and obtain the dose results through the software interface. Agreement within 9% between CTDI measurements and simulations allowed the validation of the MC program. Additionally, the algorithm for dose reporting in SimDoseCT was validated by comparing dose results from this tool with those obtained from MC simulations for three volumetric acquisitions (head, thorax and abdomen). The comparison was repeated using eight different collimations and also for another collimation in a helical abdomen examination. The results showed differences of 0.1 mSv or less for absolute dose in most organs and also in the effective dose calculation. The software provides a suitable tool for dose assessment in standard adult patients undergoing CT

  14. Study of effective dose of various protocols in equipment cone beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Soares, M. R.; Maia, A. F. [Universidade Federale de Sergipe, Departamento de Fisica, Cidade Universitaria Prof. Jose Aloisio de Campos, Marechal Rondon s/n, Jardim Rosa Elze, 49-100000 Sao Cristovao, Sergipe (Brazil); Batista, W. O. [Instituto Federal da Bahia, Rua Emidio dos Santos s/n, Barbalho, Salvador, 40301015 Bahia (Brazil); Caldas, L. V. E.; Lara, P. A., E-mail: mrs2206@gmail.com [Instituto de Pesquisas Energeticas e Nucleares / CNEN, Av. Lineu Prestes 2242, Cidade Universitaria, 05508-000 Sao Paulo (Brazil)

    2014-08-15

    Currently the cone beam computed tomography is widely used in various procedures of dental radiology. Although the doses values associated with the procedures of cone beam CT are low compared to typical values associated with dental radiology procedure in multi slices CT. However can be high compared to typical values of other techniques commonly used in dental radiology. The present scenario is a very wide range of designs of equipment and, consequently, lack of uniformity in all parameters associated with x-ray generation and geometry. In this context, this study aimed to evaluate and calculate the absorbed dose in organs and tissues relevant and estimate effective dose for different protocols with different geometries of exposure in five cone beam CT equipment. For this, a female Alderson anthropomorphic phantom, manufactured by Radiology Support Devices was used. The phantom was irradiated with 26 dosimeters LiF: Mg, Ti (TLD-100), inserted in organs and tissues along the layers forming the head and neck of the phantom. The equipment used, in this present assessment, was: i-CAT Classical, Kodak 9000 3D, Gendex GXCB 500, Sirona Orthophos X G 3D and Planmeca Pro Max 3D. The effective doses were be determined by the ICRP 103 weighting factors. The values were between 7.0 and 111.5 micro Sv, confirming the broad dose range expected due to the diversity of equipment and protocols used in each equipment. The values of effective dose per Fov size were: between 7 and 51.2 micro Sv for located Fov; between 17.6 and 52.0 micro Sv for medium Fov; and between 11.5 and 43.1 micro Sv to large Fov (maxillofacial). In obtaining the effective dose the measurements highlighted a relevance contribution of dose absorbed by the remaining organs (36%), Salivary glands (30%), thyroid (12%) and bone marrow (12%). (Author)

  15. Study of effective dose of various protocols in equipment cone beam CT

    International Nuclear Information System (INIS)

    Soares, M. R.; Maia, A. F.; Batista, W. O.; Caldas, L. V. E.; Lara, P. A.

    2014-08-01

    Currently the cone beam computed tomography is widely used in various procedures of dental radiology. Although the doses values associated with the procedures of cone beam CT are low compared to typical values associated with dental radiology procedure in multi slices CT. However can be high compared to typical values of other techniques commonly used in dental radiology. The present scenario is a very wide range of designs of equipment and, consequently, lack of uniformity in all parameters associated with x-ray generation and geometry. In this context, this study aimed to evaluate and calculate the absorbed dose in organs and tissues relevant and estimate effective dose for different protocols with different geometries of exposure in five cone beam CT equipment. For this, a female Alderson anthropomorphic phantom, manufactured by Radiology Support Devices was used. The phantom was irradiated with 26 dosimeters LiF: Mg, Ti (TLD-100), inserted in organs and tissues along the layers forming the head and neck of the phantom. The equipment used, in this present assessment, was: i-CAT Classical, Kodak 9000 3D, Gendex GXCB 500, Sirona Orthophos X G 3D and Planmeca Pro Max 3D. The effective doses were be determined by the ICRP 103 weighting factors. The values were between 7.0 and 111.5 micro Sv, confirming the broad dose range expected due to the diversity of equipment and protocols used in each equipment. The values of effective dose per Fov size were: between 7 and 51.2 micro Sv for located Fov; between 17.6 and 52.0 micro Sv for medium Fov; and between 11.5 and 43.1 micro Sv to large Fov (maxillofacial). In obtaining the effective dose the measurements highlighted a relevance contribution of dose absorbed by the remaining organs (36%), Salivary glands (30%), thyroid (12%) and bone marrow (12%). (Author)

  16. Magnitude and effects of X-ray scatter of a cone-beam micro-CT for small animal imaging

    Energy Technology Data Exchange (ETDEWEB)

    Ni, Y.C. [Institute of Nuclear Energy Research, Longtan 32546, Taiwan (China); Jan, M.L. [Institute of Nuclear Energy Research, Longtan 32546, Taiwan (China); Chen, K.W. [Institute of Nuclear Energy Research, Longtan 32546, Taiwan (China); Cheng, Y.D. [Department of Nuclear Science, National Tsing-Hua University, Hsinchu 30043, Taiwan (China); Chuang, K.S. [Department of Nuclear Science, National Tsing-Hua University, Hsinchu 30043, Taiwan (China); Fu, Y.K. [Institute of Nuclear Energy Research, Longtan 32546, Taiwan (China)]. E-mail: fufrank@iner.gov.tw

    2006-12-20

    We have developed a micro-CT system to provide high-resolution and anatomic information to combine with a microPET'' (registered) R4 system. This study was to evaluate the magnitude and effects of scatter for low kVp X-ray in this cone-beam micro-CT system. Slit collimators were used to simulate fan-beam micro-CT for comparison. The magnitudes of X-ray scatter were measured using the beam-stop method and were estimated by polynomial-fitting extrapolation to 0 mm size of stoppers. The scatter-to-primary ratio at center of the cone-beam system were 45% and 20% for rat and mouse phantoms, respectively, and were reduced to 5.86% and 4.2% in fan-beam geometric setup. The effects of X-ray scatter on image uniformity and contrast ratio were evaluated also. The uniformity response was examined by the profile of the reconstructed image. The degrees of 'cupping' in the fan-beam and cone-beam conditions were 1.75% and 3.81%, respectively, in rat phantom. A contrast phantom consisting of four inserts with physical densities similar to that of acrylic was used for measuring the effect of X-ray scatter on image contrast. Contrast ratios of the inserts and acrylic in cone-beam setup degraded 36.9% in average compared with fan-beam setup. A tumor-bearing mouse was scanned by the micro-CT system. The tumor-to-background contrast ratios were measured to be 0.331 and 0.249, respectively, with fan-beam and cone-beam setups.

  17. CT technology of which the application spreads

    International Nuclear Information System (INIS)

    Fujii, Shoji

    1990-01-01

    Computer tomography (CT) is bringing about rapid progress in measurement, analysis and nondestructive inspection techniques as it can express accurately the section of objects. As is generally known, CT technology began from the X-ray CT for human bodies which images the distribution of X-ray absorption coefficient, and developed to emission CT and magnetic resonance imaging which is harmless to living bodies. Recently in industries, the CT using X-ray and gamma ray has begun to be practically used on full scale. Also research is in progress aiming at practical use on the neutron CT using neutron beam which penetrates metals well, emission CT and ultrasonic CT. It was tried to analyze the structure of μm size and elements by the CT using synchrotron orbital radiation. The CT by serial production using 160, 320 and 420 kV X-ray tubes has begun to spread. Detectors, central controlling systems including computers, scanner mechanisms and so on are made in common, and X-ray energy can be selected conforming to scanned objects. These are used in automobile industry for the detection of defects, the accurate measurement of dimensions and so on. The spread of X-ray CT and the application of portable CT, neutron CT and automobile tire CT are reported. (K.I.)

  18. Tilted plane Feldkamp type reconstruction algorithm for spiral cone beam CT

    International Nuclear Information System (INIS)

    Yan Ming; Zhang Cishen

    2005-01-01

    An approximate image reconstruction method for spiral cone beam computed tomography (CT), called tilted plane Feldkamp type reconstruction algorithm (TPFR), is presented in this paper, which extends Feldkamp cone beam reconstruction algorithm to deal with its inaccuracy and artifact problems caused by large cone angle. This is done by tilting the reconstructing planes to minimize the cone angle and optimally fit the spiral segment of the source. The tilted plane image reconstruction requires reforming the three-dimensional projection data set for the tilted plane and application of Feldkamp algorithm to the reformed data set. Analytical and computational results can show that the image reconstruction performance of the proposed TPFR algorithm is superior to that of the Feldkamp reconstruction algorithm in the image quality, volume coverage speed, maximum achievable pitch value, and slice sensitivity profiles. Moreover, it provides more accurate image reconstruction than the existing two-dimensional reconstruction algorithms

  19. Ultrafast cone-beam CT scatter correction with GPU-based Monte Carlo simulation

    Directory of Open Access Journals (Sweden)

    Yuan Xu

    2014-03-01

    Full Text Available Purpose: Scatter artifacts severely degrade image quality of cone-beam CT (CBCT. We present an ultrafast scatter correction framework by using GPU-based Monte Carlo (MC simulation and prior patient CT image, aiming at automatically finish the whole process including both scatter correction and reconstruction within 30 seconds.Methods: The method consists of six steps: 1 FDK reconstruction using raw projection data; 2 Rigid Registration of planning CT to the FDK results; 3 MC scatter calculation at sparse view angles using the planning CT; 4 Interpolation of the calculated scatter signals to other angles; 5 Removal of scatter from the raw projections; 6 FDK reconstruction using the scatter-corrected projections. In addition to using GPU to accelerate MC photon simulations, we also use a small number of photons and a down-sampled CT image in simulation to further reduce computation time. A novel denoising algorithm is used to eliminate MC noise from the simulated scatter images caused by low photon numbers. The method is validated on one simulated head-and-neck case with 364 projection angles.Results: We have examined variation of the scatter signal among projection angles using Fourier analysis. It is found that scatter images at 31 angles are sufficient to restore those at all angles with < 0.1% error. For the simulated patient case with a resolution of 512 × 512 × 100, we simulated 5 × 106 photons per angle. The total computation time is 20.52 seconds on a Nvidia GTX Titan GPU, and the time at each step is 2.53, 0.64, 14.78, 0.13, 0.19, and 2.25 seconds, respectively. The scatter-induced shading/cupping artifacts are substantially reduced, and the average HU error of a region-of-interest is reduced from 75.9 to 19.0 HU.Conclusion: A practical ultrafast MC-based CBCT scatter correction scheme is developed. It accomplished the whole procedure of scatter correction and reconstruction within 30 seconds.----------------------------Cite this

  20. Beam-hardening correction in CT based on basis image and TV model

    International Nuclear Information System (INIS)

    Li Qingliang; Yan Bin; Li Lei; Sun Hongsheng; Zhang Feng

    2012-01-01

    In X-ray computed tomography, the beam hardening leads to artifacts and reduces the image quality. It analyzes how beam hardening influences on original projection. According, it puts forward a kind of new beam-hardening correction method based on the basis images and TV model. Firstly, according to physical characteristics of the beam hardening an preliminary correction model with adjustable parameters is set up. Secondly, using different parameters, original projections are operated by the correction model. Thirdly, the projections are reconstructed to obtain a series of basis images. Finally, the linear combination of basis images is the final reconstruction image. Here, with total variation for the final reconstruction image as the cost function, the linear combination coefficients for the basis images are determined according to iterative method. To verify the effectiveness of the proposed method, the experiments are carried out on real phantom and industrial part. The results show that the algorithm significantly inhibits cup and strip artifacts in CT image. (authors)

  1. An efficient CT-simulation procedure for breast treatment using tangent beams

    International Nuclear Information System (INIS)

    Lu, H.-M.; Cheng Pan; Lee, Chin; Svensson, Goran; Harris, Jay

    1997-01-01

    Purpose:Breast treatment planning using CT-simulations provides a number of advantages, but presents several unique problems. One concern is the ability to evaluate coverage of the external target volume, since CT scanners cannot provide field light projections on skin. Another is whether treatment portals can be marked on a patient as in a regular simulation, so that the usual level of setup accuracy can be achieved without additional effort at the treatment unit. Finally, the planning procedure must be performed with efficiency, so that it could be used routinely for most or all patients. To address these issues, we report our CT-simulation procedure for breast treatment using tangent beams, which includes, all in one session, CT data acquisition, field determination, external target volume evaluation, and the marking of treatment portals on the patient. Methods and Materials: A General Electric CT scanner and a virtual simulation software package 'Advantage-Sim' were used to perform the simulations. We have developed two tools to assist the simulation process. One is a digitization system consisting of multimedia software interacting with a sonic digitizer that can capture the coordinates of a point in space with the standard deviation of 1.4 mm. Given the planned beam geometry, the system can establish a virtual beam projection on the patient's body in real space, so that for any digitized point on the skin, its position relative to the field can be calculated and shown in a beam's eye view display. With convenient audio and visual signals, the system allows one to see if any skin area of concern is included in the field with sufficient margin, or to rapidly locate field borders or marking points by cruising the digitizer probe on patient skin. The accuracy of the system has been studied by using a breast phantom. The other is a breast planning software tool which augments the virtual simulation software to speed up the generation of tangent beam pairs

  2. The Relationships of the Maxillary Sinus With the Superior Alveolar Nerves and Vessels as Demonstrated by Cone-Beam CT Combined With μ-CT and Histological Analyses.

    Science.gov (United States)

    Kasahara, Norio; Morita, Wataru; Tanaka, Ray; Hayashi, Takafumi; Kenmotsu, Shinichi; Ohshima, Hayato

    2016-05-01

    There are no available detailed data on the three-dimensional courses of the human superior alveolar nerves and vessels. This study aimed to clarify the relationships of the maxillary sinus with the superior alveolar nerves and vessels using cone-beam computed tomography (CT) combined with μ-CT and histological analyses. Digital imaging and communication in medicine data obtained from the scanned heads/maxillae of cadavers used for undergraduate/postgraduate dissection practice and skulls using cone-beam CT were reconstructed into three-dimensional (3D) images using software. The 3D images were compared with μ-CT images and histological sections. Cone-beam CT clarified the relationships of the maxillary sinus with the superior alveolar canals/grooves. The main anterior superior alveolar canal/groove ran anteriorly through the upper part of the sinus and terminated at the bottom of the nasal cavity near the piriform aperture. The main middle alveolar canal ran downward from the upper part of the sinus to ultimately join the anterior one. The main posterior alveolar canal ran through the lateral lower part of the sinus and communicated with the anterior one. Histological analyses demonstrated the existence of nerves and vessels in these canals/grooves, and the quantities of these structures varied across each canal/groove. Furthermore, the superior dental nerve plexus exhibited a network that was located horizontally to the occlusal plane, although these nerve plexuses appeared to be the vertical network that is described in most textbooks. In conclusion, cone-beam CT is suggested to be a useful method for clarifying the superior alveolar canals/grooves including the nerves and vessels. © 2016 Wiley Periodicals, Inc.

  3. Dose optimisation for intraoperative cone-beam flat-detector CT in paediatric spinal surgery

    International Nuclear Information System (INIS)

    Petersen, Asger Greval; Eiskjaer, Soeren; Kaspersen, Jon

    2012-01-01

    During surgery for spinal deformities, accurate placement of pedicle screws may be guided by intraoperative cone-beam flat-detector CT. The purpose of this study was to identify appropriate paediatric imaging protocols aiming to reduce the radiation dose in line with the ALARA principle. Using O-arm registered (Medtronic, Inc.), three paediatric phantoms were employed to measure CTDI w doses with default and lowered exposure settings. Images from 126 scans were evaluated by two spinal surgeons and scores were compared (Kappa statistics). Effective doses were calculated. The recommended new low-dose 3-D spine protocols were then used in 15 children. The lowest acceptable exposure as judged by image quality for intraoperative use was 70 kVp/40 mAs, 70 kVp/80 mAs and 80 kVp/40 mAs for the 1-, 5- and 12-year-old-equivalent phantoms respectively (kappa = 0,70). Optimised dose settings reduced CTDI w doses 89-93%. The effective dose was 0.5 mSv (91-94,5% reduction). The optimised protocols were used clinically without problems. Radiation doses for intraoperative 3-D CT using a cone-beam flat-detector scanner could be reduced at least 89% compared to manufacturer settings and still be used to safely navigate pedicle screws. (orig.)

  4. Tooth labeling in cone-beam CT using deep convolutional neural network for forensic identification

    Science.gov (United States)

    Miki, Yuma; Muramatsu, Chisako; Hayashi, Tatsuro; Zhou, Xiangrong; Hara, Takeshi; Katsumata, Akitoshi; Fujita, Hiroshi

    2017-03-01

    In large disasters, dental record plays an important role in forensic identification. However, filing dental charts for corpses is not an easy task for general dentists. Moreover, it is laborious and time-consuming work in cases of large scale disasters. We have been investigating a tooth labeling method on dental cone-beam CT images for the purpose of automatic filing of dental charts. In our method, individual tooth in CT images are detected and classified into seven tooth types using deep convolutional neural network. We employed the fully convolutional network using AlexNet architecture for detecting each tooth and applied our previous method using regular AlexNet for classifying the detected teeth into 7 tooth types. From 52 CT volumes obtained by two imaging systems, five images each were randomly selected as test data, and the remaining 42 cases were used as training data. The result showed the tooth detection accuracy of 77.4% with the average false detection of 5.8 per image. The result indicates the potential utility of the proposed method for automatic recording of dental information.

  5. SU-E-J-103: Setup Errors Analysis by Cone-Beam CT (CBCT)-Based Imaged-Guided Intensity Modulated Radiotherapy for Esophageal Cancer

    Energy Technology Data Exchange (ETDEWEB)

    Yang, H; Wang, W; Hu, W; Chen, X; Wang, X; Yu, C [Taizhou Hospital, Wenzhou Medical College, Taizhou, Zhejiang (China)

    2014-06-01

    Purpose: To quantify setup errors by pretreatment kilovolt cone-beam computed tomography(KV-CBCT) scans for middle or distal esophageal carcinoma patients. Methods: Fifty-two consecutive middle or distal esophageal carcinoma patients who underwent IMRT were included this study. A planning CT scan using a big-bore CT simulator was performed in the treatment position and was used as the reference scan for image registration with CBCT. CBCT scans(On-Board Imaging v1. 5 system, Varian Medical Systems) were acquired daily during the first treatment week. A total of 260 CBCT scans was assessed with a registration clip box defined around the PTV-thorax in the reference scan based on(nine CBCTs per patient) bony anatomy using Offline Review software v10.0(Varian Medical Systems). The anterior-posterior(AP), left-right(LR), superiorinferior( SI) corrections were recorded. The systematic and random errors were calculated. The CTV-to-PTV margins in each CBCT frequency was based on the Van Herk formula (2.5Σ+0.7σ). Results: The SD of systematic error (Σ) was 2.0mm, 2.3mm, 3.8mm in the AP, LR and SI directions, respectively. The average random error (σ) was 1.6mm, 2.4mm, 4.1mm in the AP, LR and SI directions, respectively. The CTV-to-PTV safety margin was 6.1mm, 7.5mm, 12.3mm in the AP, LR and SI directions based on van Herk formula. Conclusion: Our data recommend the use of 6 mm, 8mm, and 12 mm for esophageal carcinoma patient setup in AP, LR, SI directions, respectively.

  6. TU-H-207A-03: CT Hounsfield Unit Accuracy: Effect of Beam Hardening On Phantom and Clinical Whole-Body CT Images

    Energy Technology Data Exchange (ETDEWEB)

    Ai, H; Wendt, R [The University of Texas MD Anderson Cancer Center, Houston, TX (United States)

    2016-06-15

    Purpose: To assess the effect of beam hardening on measured CT HU values. Methods: An anthropomorphic knee phantom was scanned with the CT component of a GE Discovery 690 PET/CT scanner (120kVp, 300mAs, 40?0.625mm collimation, pitch=0.984, FOV=500mm, matrix=512?512) with four different scan setups, each of which induces different degrees of beam hardening by introducing additional attenuation media into the field of view. Homogeneous voxels representing “soft tissue” and “bone” were segmented by HU thresholding followed by a 3D morphological erosion operation which removes the non-homogenous voxels located on the interface of thresholded tissue mask. HU values of segmented “soft tissue” and “bone” were compared.Additionally, whole-body CT data with coverage from the skull apex to the end of toes were retrospectively retrieved from seven PET/CT exams to evaluate the effect of beam hardening in vivo. Homogeneous bone voxels were segmented with the same method previously described. Total In-Slice Attenuation (TISA) for each CT slice, defined as the summation of HU values over all voxels within a CT slice, was calculated for all slices of the seven whole-body CT datasets and evaluated against the mean HU values of homogeneous bone voxels within that slice. Results: HU values measured from the phantom showed that while “soft tissue” HU values were unaffected, added attenuation within the FOV caused noticeable decreases in the measured HU values of “bone” voxels. A linear relationship was observed between bone HU and TISA for slices of the torso and legs, but not of the skull. Conclusion: Beam hardening effect is not an issue of concern for voxels with HU in the soft tissue range, but should not be neglected for bone voxels. A linear relationship exists between bone HU and the associated TISA in non-skull CT slices, which can be exploited to develop a correction strategy.

  7. The assessment of trabecular bone parameters and cortical bone strength: a comparison of micro-CT and dental cone-beam CT.

    Science.gov (United States)

    Hsu, Jui-Ting; Wang, Shun-Ping; Huang, Heng-Li; Chen, Ying-Ju; Wu, Jay; Tsai, Ming-Tzu

    2013-10-18

    This study compared the capabilities of micro-computed tomography (micro-CT) and dental cone-beam computed tomography (CBCT) in assessing trabecular bone parameters and cortical bone strength. Micro-CT and CBCT scans were applied to 28 femurs from 14 rats to obtain independent measurements of the volumetric cancellous bone mineral density (vCanBMD) in the femoral head, volumetric cortical bone mineral density (vCtBMD) in the femoral diaphysis, cross-sectional moment of inertia (CSMI), and bone strength index (BSI) (=CSMI×vCtBMD). Five structural parameters of the trabecular bone of the femoral head were calculated from micro-CT images. A three-point bending test was then conducted to measure the fracture load of each femur. Bivariate linear Pearson analysis was conducted to calculate the correlation coefficients (r values) of the micro-CT, dental CBCT, and three-point bending measurements. The statistical analyses showed a strong correlation between vCanBMD values obtained using micro-CT and dental CBCT (r=0.830). There were strong or moderate correlation between vCanBMD measured using dental CBCT and five parameters of trabecular structure measured using micro-CT. Additionally, the results were satisfactory regardless of whether micro-CT or dental CBCT was used to measure the femoral diaphysis vCtBMD (r=0.733 and 0.680, respectively), CSMI (r=0.756 and 0.726, respectively), or BSI (r=0.846 and 0.847, respectively) to predict fracture loads. This study has yielded a new method for using dental CBCT to evaluate bone parameters and bone strength; however, further studies are necessary to validate the use of dental CBCT on humans. © 2013 Elsevier Ltd. All rights reserved.

  8. Reduction of Cone-Beam CT scan time without compromising the accuracy of the image registration in IGRT

    DEFF Research Database (Denmark)

    Westberg, Jonas; Jensen, Henrik R; Bertelsen, Anders

    2010-01-01

    In modern radiotherapy accelerators are equipped with 3D cone-beam CT (CBCT) which is used to verify patient position before treatment. The verification is based on an image registration between the CBCT acquired just before treatment and the CT scan made for the treatment planning. The purpose...... of this study is to minimise the scan time of the CBCT without compromising the accuracy of the image registration in IGRT....

  9. The Role of Cone-Beam CT in Transcatheter Arterial Chemoembolization for Hepatocellular Carcinoma: A Systematic Review and Meta-analysis.

    Science.gov (United States)

    Pung, Leland; Ahmad, Moiz; Mueller, Kerstin; Rosenberg, Jarrett; Stave, Christopher; Hwang, Gloria L; Shah, Rajesh; Kothary, Nishita

    2017-03-01

    To review available evidence for use of cone-beam CT during transcatheter arterial chemoembolization in hepatocellular carcinoma (HCC) for detection of tumor and feeding arteries. Literature searches were conducted from inception to May 15, 2016, in PubMed (MEDLINE), Scopus, and Cochrane Central Register of Controlled Trials. Searches included "cone beam," "CBCT," "C-arm," "CACT," "cone-beam CT," "volumetric CT," "volume computed tomography," "volume CT," AND "liver," "hepatic*," "hepatoc*." Studies that involved adults with HCC specifically and treated with transcatheter arterial chemoembolization that used cone-beam CT were included. Inclusion criteria were met by 18 studies. Pooled sensitivity of cone-beam CT for detecting tumor was 90% (95% confidence interval [CI], 82%-95%), whereas pooled sensitivity of digital subtraction angiography (DSA) for tumor detection was 67% (95% CI, 51%-80%). Pooled sensitivity of cone-beam CT for detecting tumor feeding arteries was 93% (95% CI, 91%-95%), whereas pooled sensitivity of DSA was 55% (95% CI, 36%-74%). Cone-beam CT can significantly increase detection of tumors and tumor feeding arteries during transcatheter arterial chemoembolization. Cone-beam CT should be considered as an adjunct tool to DSA during transcatheter arterial chemoembolization treatments of HCC. Copyright © 2016 SIR. Published by Elsevier Inc. All rights reserved.

  10. Investigation of C-arm cone-beam CT-guided surgery of the frontal recess.

    Science.gov (United States)

    Rafferty, M A; Siewerdsen, J H; Chan, Y; Moseley, D J; Daly, M J; Jaffray, D A; Irish, J C

    2005-12-01

    A cone-beam CT (CBCT) imaging system based on a mobile C-arm (Siemens PowerMobil) incorporating a high-performance flat-panel detector (Varian PaxScan) has been developed in our laboratory. We hypothesize that intraoperative C-arm CBCT provides image quality and guidance performance sufficient to assist surgical approach to the frontal recess. A preclinical prospective study was conducted using six cadaver heads to assess the performance characteristics and the potential clinical utility of this imaging system. The mobile C-arm was employed for intraoperative CBCT guidance of the endoscopic approach to twelve frontal recesses. The imaging system is capable of sub-mm 3D spatial resolution with bone and soft-tissue visibility and a field of view sufficient for guidance of head and neck surgery. The system can generate intraoperative, volumetric CT images rapidly with an acceptably low radiation exposure to the patient and with image quality sufficient for most surgical tasks. Moreover, the system is portable and compatible with the surgical setup, providing excellent access to the patient. Finally, the accuracy of the system is not bound to a registration process. The ability to create updated images as surgery progresses introduces the concept of 'near-real-time' CT guidance for head and neck surgery. We found that the use of CBCT increased surgical confidence in accessing the frontal recess, resolved ambiguities with anatomical variations, and provided valuable teaching information to surgeons in training in both preoperative planning and correlation between tri-planar CT scans and intraoperative endoscopic findings.

  11. Detection of experimental acute myocardial infarction in dogs with electron beam CT

    International Nuclear Information System (INIS)

    Zhang Minguang; Song Zhijian; Shen Tianzhen; Zuo Huansen; Feng Xiaoyuan; Chen Xingrong

    1999-01-01

    Objective: The value and limitation of discovering the site, shape, density and size of acute myocardial infarction with enhanced volume scans of electron beam computed tomography (EBCT) in experimental dogs were assessed. Methods: The anterior descending and circumflex branches of the left coronary artery were ligated and infarction in the wall of left ventricle was created in six dogs. Enhancement volume scan was made with a EBCT scanner at 1.5, 6, 12, and 20 hour after the ligation and CT sectional images of dog heart were obtained. After finishing the scanning, the dogs were then scarified, and the samples were sent for electron and optic microscopes for pathologic examination. The morphologic findings of myocardial ischemia and infarction in CT images were carefully analysed. The CT numbers and the ratios of left ventricular myocardial and luminal CT number (M/L) in the myocardial ischemic and infarcted areas and normal areas were calculated. Results: The typical myocardial infarction in areas ligated were confirmed by ECG and pathology 20 hours after ligation. The site and shape of acute myocardial infarction were clearly shown on the EBCT images. The density of the ischemia and infarction area was significantly lower than that of the normal myocardium (P < 0.001). M/Ls of the low density areas and normal areas were 13.2% and 32.9%, respectively. The difference was significant (P < 0.001). Conclusions: With adequate scanning technique, the site shape, density and size of acute myocardial ischemia and infarction can be accurately assessed by EBCT

  12. Hounsfield unit recovery in clinical cone beam CT images of the thorax acquired for image guided radiation therapy

    DEFF Research Database (Denmark)

    Thing, Rune Slot; Bernchou, Uffe; Mainegra-Hing, Ernesto

    2016-01-01

    A comprehensive artefact correction method for clinical cone beam CT (CBCT) images acquired for image guided radiation therapy (IGRT) on a commercial system is presented. The method is demonstrated to reduce artefacts and recover CT-like Hounsfield units (HU) in reconstructed CBCT images of five...... lung cancer patients. Projection image based artefact corrections of image lag, detector scatter, body scatter and beam hardening are described and applied to CBCT images of five lung cancer patients. Image quality is evaluated through visual appearance of the reconstructed images, HU...

  13. SU-G-JeP3-06: Lower KV Image Dose Are Expected From a Limited-Angle Intra-Fractional Verification (LIVE) System for SBRT Treatments

    International Nuclear Information System (INIS)

    Ding, G; Yin, F; Ren, L

    2016-01-01

    Purpose: In order to track the tumor movement for patient positioning verification during arc treatment delivery or in between 3D/IMRT beams for stereotactic body radiation therapy (SBRT), the limited-angle kV projections acquisition simultaneously during arc treatment delivery or in-between static treatment beams as the gantry moves to the next beam angle was proposed. The purpose of this study is to estimate additional imaging dose resulting from multiple tomosynthesis acquisitions in-between static treatment beams and to compare with that of a conventional kV-CBCT acquisition. Methods: kV imaging system integrated into Varian TrueBeam accelerators was modeled using EGSnrc Monte Carlo user code, BEAMnrc and DOSXYZnrc code was used in dose calculations. The simulated realistic kV beams from the Varian TrueBeam OBI 1.5 system were used to calculate dose to patient based on CT images. Organ doses were analyzed using DVHs. The imaging dose to patient resulting from realistic multiple tomosynthesis acquisitions with each 25–30 degree kV source rotation between 6 treatment beam gantry angles was studied. Results: For a typical lung SBRT treatment delivery much lower (20–50%) kV imaging doses from the sum of realistic six tomosynthesis acquisitions with each 25–30 degree x-ray source rotation between six treatment beam gantry angles were observed compared to that from a single CBCT image acquisition. Conclusion: This work indicates that the kV imaging in this proposed Limited-angle Intra-fractional Verification (LIVE) System for SBRT Treatments has a negligible imaging dose increase. It is worth to note that the MV imaging dose caused by MV projection acquisition in-between static beams in LIVE can be minimized by restricting the imaging to the target region and reducing the number of projections acquired. For arc treatments, MV imaging acquisition in LIVE does not add additional imaging dose as the MV images are acquired from treatment beams directly during the

  14. SU-F-J-109: Generate Synthetic CT From Cone Beam CT for CBCT-Based Dose Calculation

    International Nuclear Information System (INIS)

    Wang, H; Barbee, D; Wang, W; Pennell, R; Hu, K; Osterman, K

    2016-01-01

    Purpose: The use of CBCT for dose calculation is limited by its HU inaccuracy from increased scatter. This study presents a method to generate synthetic CT images from CBCT data by a probabilistic classification that may be robust to CBCT noise. The feasibility of using the synthetic CT for dose calculation is evaluated in IMRT for unilateral H&N cancer. Methods: In the training phase, a fuzzy c-means classification was performed on HU vectors (CBCT, CT) of planning CT and registered day-1 CBCT image pair. Using the resulting centroid CBCT and CT values for five classified “tissue” types, a synthetic CT for a daily CBCT was created by classifying each CBCT voxel to obtain its probability belonging to each tissue class, then assigning a CT HU with a probability-weighted summation of the classes’ CT centroids. Two synthetic CTs from a CBCT were generated: s-CT using the centroids from classification of individual patient CBCT/CT data; s2-CT using the same centroids for all patients to investigate the applicability of group-based centroids. IMRT dose calculations for five patients were performed on the synthetic CTs and compared with CT-planning doses by dose-volume statistics. Results: DVH curves of PTVs and critical organs calculated on s-CT and s2-CT agree with those from planning-CT within 3%, while doses calculated with heterogeneity off or on raw CBCT show DVH differences up to 15%. The differences in PTV D95% and spinal cord max are 0.6±0.6% and 0.6±0.3% for s-CT, and 1.6±1.7% and 1.9±1.7% for s2-CT. Gamma analysis (2%/2mm) shows 97.5±1.6% and 97.6±1.6% pass rates for using s-CTs and s2-CTs compared with CT-based doses, respectively. Conclusion: CBCT-synthesized CTs using individual or group-based centroids resulted in dose calculations that are comparable to CT-planning dose for unilateral H&N cancer. The method may provide a tool for accurate dose calculation based on daily CBCT.

  15. SU-F-J-109: Generate Synthetic CT From Cone Beam CT for CBCT-Based Dose Calculation

    Energy Technology Data Exchange (ETDEWEB)

    Wang, H; Barbee, D; Wang, W; Pennell, R; Hu, K; Osterman, K [Department of Radiation Oncology, NYU Langone Medical Center, New York, NY (United States)

    2016-06-15

    Purpose: The use of CBCT for dose calculation is limited by its HU inaccuracy from increased scatter. This study presents a method to generate synthetic CT images from CBCT data by a probabilistic classification that may be robust to CBCT noise. The feasibility of using the synthetic CT for dose calculation is evaluated in IMRT for unilateral H&N cancer. Methods: In the training phase, a fuzzy c-means classification was performed on HU vectors (CBCT, CT) of planning CT and registered day-1 CBCT image pair. Using the resulting centroid CBCT and CT values for five classified “tissue” types, a synthetic CT for a daily CBCT was created by classifying each CBCT voxel to obtain its probability belonging to each tissue class, then assigning a CT HU with a probability-weighted summation of the classes’ CT centroids. Two synthetic CTs from a CBCT were generated: s-CT using the centroids from classification of individual patient CBCT/CT data; s2-CT using the same centroids for all patients to investigate the applicability of group-based centroids. IMRT dose calculations for five patients were performed on the synthetic CTs and compared with CT-planning doses by dose-volume statistics. Results: DVH curves of PTVs and critical organs calculated on s-CT and s2-CT agree with those from planning-CT within 3%, while doses calculated with heterogeneity off or on raw CBCT show DVH differences up to 15%. The differences in PTV D95% and spinal cord max are 0.6±0.6% and 0.6±0.3% for s-CT, and 1.6±1.7% and 1.9±1.7% for s2-CT. Gamma analysis (2%/2mm) shows 97.5±1.6% and 97.6±1.6% pass rates for using s-CTs and s2-CTs compared with CT-based doses, respectively. Conclusion: CBCT-synthesized CTs using individual or group-based centroids resulted in dose calculations that are comparable to CT-planning dose for unilateral H&N cancer. The method may provide a tool for accurate dose calculation based on daily CBCT.

  16. USING A 100 KV DC LOAD LOCK PHOTOGUN TO MEASURE PHOTOCATHODE LIFETIME OF HIGH POLARIZATION STRAINED SUPERLATTICE GAAS/GAASP AT BEAM INTENSITY >1 MILLIAMP

    International Nuclear Information System (INIS)

    Joseph Grames; Benard Poelker; Philip Adderley; Joshua Brittian; James Clark; John Hansknecht; Danny Machie; Marcy Stutzman; Kenneth Surles-law; Riad Suleiman

    2007-01-01

    A new GaAs DC high voltage load lock photogun has been constructed at Jefferson Laboratory (JLab), with improved vacuum and photocathode preparation capabilities. As reported previously, this gun was used to study photocathode lifetime with bulk GaAs at DC beam currents between 1 and 10 mA. In this submission, lifetime measurements were performed using high polarization strained-superlattice GaAs photocathode material at beam currents up to 1 mA, with near bandgap light from a fiber based drive laser having picosecond optical pulses and RF time structure

  17. Empirical binary tomography calibration (EBTC) for the precorrection of beam hardening and scatter for flat panel CT.

    Science.gov (United States)

    Grimmer, Rainer; Kachelriess, Marc

    2011-04-01

    Scatter and beam hardening are prominent artifacts in x-ray CT. Currently, there is no precorrection method that inherently accounts for tube voltage modulation and shaped prefiltration. A method for self-calibration based on binary tomography of homogeneous objects, which was proposed by B. Li et al. ["A novel beam hardening correction method for computed tomography," in Proceedings of the IEEE/ICME International Conference on Complex Medical Engineering CME 2007, pp. 891-895, 23-27 May 2007], has been generalized in order to use this information to preprocess scans of other, nonbinary objects, e.g., to reduce artifacts in medical CT applications. Further on, the method was extended to handle scatter besides beam hardening and to allow for detector pixel-specific and ray-specific precorrections. This implies that the empirical binary tomography calibration (EBTC) technique is sensitive to spectral effects as they are induced by the heel effect, by shaped prefiltration, or by scanners with tube voltage modulation. The presented method models the beam hardening correction by using a rational function, while the scatter component is modeled using the pep model of B. Ohnesorge et al. ["Efficient object scatter correction algorithm for third and fourth generation CT scanners," Eur. Radiol. 9(3), 563-569 (1999)]. A smoothness constraint is applied to the parameter space to regularize the underdetermined system of nonlinear equations. The parameters determined are then used to precorrect CT scans. EBTC was evaluated using simulated data of a flat panel cone-beam CT scanner with tube voltage modulation and bow-tie prefiltration and using real data of a flat panel cone-beam CT scanner. In simulation studies, where the ground truth is known, the authors' correction model proved to be highly accurate and was able to reduce beam hardening by 97% and scatter by about 75%. Reconstructions of measured data showed significantly less artifacts than the standard reconstruction

  18. Using condition and usefulness of dental cone-beam CT in endodontic treatment

    International Nuclear Information System (INIS)

    Kimura, Yuichi; Araki, Kazuyuki; Yamada, Yoshishige; Tagaya, Atsuko; Seki, Kenji; Okano, Tomohiro; Endo, Atsushi

    2009-01-01

    This study evaluated the condition and usefulness of the dental cone-beam CT (3DX) in clinical endodontic treatments. Images from 55 examinations of 49 patients obtained using 3DX during an 11-month period were evaluated retrospectively to identify the usefulness of this modality compared with periapical or panoramic radiographs. The main indication for using of 3DX was diagnosis of root fracture in 65% of the examinations, second was the presence and expansion of periapical lesion in 22%, and third was to detect the canal system or root abnormality in 13%. The 3DX visualizes bony anatomical structures precisely and detects the presence and expansion of periapical lesions and the canal system of each root of mulirooted teeth that cannot easily be observed by intraoral radiography or panoramic radiography. The results of this study suggest that 3DX is a useful and reliable tool for endodontic treatments. (author)

  19. MicroCT Bone Densitometry: Context Sensitivity, Beam Hardening Correction and the Effect of Surrounding Media

    Directory of Open Access Journals (Sweden)

    Philip L. Salmon

    2014-12-01

    Full Text Available The context-sensitivity of microCT bone densitometry due to beam hardening artefacts was assessed. Bones and teeth are scanned with varying thickness of surrounding media (water, alcohol, biological tissue and it is important to understand how this affects reconstructed attenuation (“density” of the mineralized tissue. Aluminium tubes and rods with thickness 0.127mm–5mm were scanned both in air or surrounded by up to 2cm of water. Scans were performed with different energy filters and degrees of software beam hardening correction (BHC. Also tested were the effects of signal-to-noise ratio, magnification and truncation. The thickness of an aluminium tube significantly affected its mean reconstructed attenuation. This effect of thickness could be reduced substantially by BHC for scans in air, but not for scans in water. Varying thickness of surrounding water also changed the mean attenuation of an aluminium tube. This artefact could be almost eliminated by an optimal BHC value. The “cupping” artefact of heterogeneous attenuation (elevated at outer surfaces could be corrected if aluminium was scanned in air, but in water BHC was much less effective. Scan truncation, changes to magnification and signal-to-noise ratio also caused artificial changes to measured attenuation. Measurement of bone mineral density by microCT is highly context sensitive. A surrounding layer of liquid or biological tissue reduces the ability of software BHC to remove bone density artefacts. Sample thickness, truncation, magnification and signal to noise ratio also affect reconstructed attenuation. Thus it is important for densitometry that sample and calibration phantom dimensions and mounting materials are standardised.

  20. Regularized iterative weighted filtered backprojection for helical cone-beam CT

    International Nuclear Information System (INIS)

    Sunnegaardh, Johan; Danielsson, Per-Erik

    2008-01-01

    Contemporary reconstruction methods employed for clinical helical cone-beam computed tomography (CT) are analytical (noniterative) but mathematically nonexact, i.e., the reconstructed image contains so called cone-beam artifacts, especially for higher cone angles. Besides cone artifacts, these methods also suffer from windmill artifacts: alternating dark and bright regions creating spiral-like patterns occurring in the vicinity of high z-direction derivatives. In this article, the authors examine the possibility to suppress cone and windmill artifacts by means of iterative application of nonexact three-dimensional filtered backprojection, where the analytical part of the reconstruction brings about accelerated convergence. Specifically, they base their investigations on the weighted filtered backprojection method [Stierstorfer et al., Phys. Med. Biol. 49, 2209-2218 (2004)]. Enhancement of high frequencies and amplification of noise is a common but unwanted side effect in many acceleration attempts. They have employed linear regularization to avoid these effects and to improve the convergence properties of the iterative scheme. Artifacts and noise, as well as spatial resolution in terms of modulation transfer functions and slice sensitivity profiles have been measured. The results show that for cone angles up to ±2.78 deg., cone artifacts are suppressed and windmill artifacts are alleviated within three iterations. Furthermore, regularization parameters controlling spatial resolution can be tuned so that image quality in terms of spatial resolution and noise is preserved. Simulations with higher number of iterations and long objects (exceeding the measured region) verify that the size of the reconstructible region is not reduced, and that the regularization greatly improves the convergence properties of the iterative scheme. Taking these results into account, and the possibilities to extend the proposed method with more accurate modeling of the acquisition process

  1. KV7 potassium channels

    DEFF Research Database (Denmark)

    Stott, Jennifer B; Jepps, Thomas Andrew; Greenwood, Iain A

    2014-01-01

    Potassium channels are key regulators of smooth muscle tone, with increases in activity resulting in hyperpolarisation of the cell membrane, which acts to oppose vasoconstriction. Several potassium channels exist within smooth muscle, but the KV7 family of voltage-gated potassium channels have been...

  2. Intraoperative cone-beam CT for guidance of temporal bone surgery.

    Science.gov (United States)

    Rafferty, Mark A; Siewerdsen, Jeffrey H; Chan, Yvonne; Daly, Michael J; Moseley, Douglas J; Jaffray, David A; Irish, Jonathan C

    2006-05-01

    To describe our preclinical experience with Cone Beam CT (CBCT) in image-guided surgery of the temporal bone. A mobile isocentric C-arm (PowerMobil, Siemens Medical Systems, Erlangen, Germany) modified to include a flat-panel detector (Varian Imaging Products, Palo Alto, CA) and a motorized orbit was developed to acquire multiple projections in rotation about a subject. Initial experiments imaging steel wire in air were used to investigate the system's spatial resolution in 3D image reconstruction. Subsequently temporal bone dissection was performed on five cadaver heads using the modified C-arm as an image guidance system. We obtained a spatial resolution of 0.85 mm. The image acquisition time was 120 seconds and the radiation dose approximately one-tenth of a conventional CT scan. CBCT provided submillimeter accuracy at high speed with low radiation dosage to offer utility as an intraoperative imaging system. CBCT offers technology that approximates "near-real-time" image guidance. C-4.

  3. Evaluation of the spatial resolution characteristics of a cone-beam breast CT scanner

    International Nuclear Information System (INIS)

    Kwan, Alexander L. C.; Boone, John M.; Yang, K.; Huang, S.-Y.

    2007-01-01

    The purpose of this study was to examine the spatial resolution of a prototype pendant-geometry cone-beam breast computed tomography (CT) system. Modulation transfer functions (MTFs) of the reconstructed image in the coronal (x and y) plane were computed as a function of the cone angle, the radial distance from the axis of rotation, the size of the reconstruction matrix, the back-projection filter used, and the number of projections acquired for the reconstruction. The results show that the cone angle and size of the reconstruction matrix have minimal impact on the MTF, while the MTF degraded radially from the axis of rotation (from 0.76 at 2.6 mm from axis of rotation down to 0.37 at 76.9 mm from axis of rotation at f=0.5 mm -1 ). The Ramp reconstruction filter increases the MTF near the axis of rotation relative to the Shepp-Logan filter, while an increase in the number of projections from 500 to 1000 increased the MTF near the periphery of the reconstructed image. The MTF in the z direction (anterior-posterior direction) was also evaluated. The z-direction MTF values tend to be higher when compared to the coronal MTF (0.85 at f=0.5 mm -1 ), and tend to be very constant throughout the coronal plane direction. The results suggest that an increase in the MTF for the prototype breast CT system is possible by optimizing various scanning and reconstruction parameters

  4. Bowtie filter and water calibration in the improvement of cone beam CT image quality

    International Nuclear Information System (INIS)

    Li Minghui; Dai Jianrong; Zhang Ke

    2010-01-01

    Objective: To evaluate the improvement of cone beam CT (CBCT) image quality by using bewtie filter (F 1 ) and water calibration. Methods: First the multi-level gain calibration of the detector panel with the method of Cal 2 calibration was performed, and the CT images of CATPHAN503 with F 0 and bowtie filter were collected, respectively. Then the detector panel using water calibration kit was calibrated, and images were acquired again. Finally, the change of image quality after using F 1 and (or) water calibration method was observed. The observed indexes included low contrast visibility, spatial uniformity, ring artifact, spatial resolution and geometric accuracy. Results: Comparing with the traditional combination of F 0 filter and Cal 2 calibration, the combination of bowtie filter F 1 and water calibration improves low contrast visibility by 13.71%, and spatial uniformity by 54. 42%. Water calibration removes ring artifacts effectively. However, none of them improves spatial resolution and geometric accuracy. Conclusions: The combination of F 1 and water calibration improves CBCT image quality effectively. This improvement is aid to the registration of CBCT images and localization images. (authors)

  5. Fixed, object-specific intensity compensation for cone beam optical CT radiation dosimetry

    Science.gov (United States)

    Dekker, Kurtis H.; Hazarika, Rubin; Silveira, Matheus A.; Jordan, Kevin J.

    2018-03-01

    Optical cone beam computed tomography (CT) scanning of radiochromic gel dosimeters, using a CCD camera and a low stray light convergent source, provides fast, truly 3D radiation dosimetry with high accuracy. However, a key limiting factor in radiochromic gel dosimetry at large (⩾10 cm diameter) volumes is the initial attenuation of the dosimeters. It is not unusual to observe a 5–10×  difference in signal intensity through the dosimeter center versus through the surrounding medium in pre-irradiation images. Thus, all dosimetric information in a typical experiment is measured within the lower 10%–20% of the camera sensor’s range, and re-use of gels is often not possible due to a lack of transmission. To counteract this, in this note we describe a simple method to create source compensators by printing on transparent films. This technique, which is easily implemented and inexpensive, is an optical analogue to the bowtie filter in x-ray CT. We present transmission images and solution phantom reconstructions to demonstrate that (1) placing compensators beyond the focal zone of the imaging lens prevents high spatial frequency features of the printed films from generating reconstruction artifacts, and (2) object-specific compensation considerably reduces the range of intensities measured in projection images. This will improve the measurable dose range in optical CT dosimetry, and will enable imaging of larger gel volumes (∼15 cm diameter). Additionally, it should enable re-use of dosimeters by printing a new compensator for a second experiment.

  6. Adaptive cone-beam CT planning improves long-term biochemical disease-free survival for 125I prostate brachytherapy

    NARCIS (Netherlands)

    Peters, M.; Smit Duijzentkunst, D. A.; Westendorp, H.; van de Pol, S. M G; Kattevilder, R.; Schellekens, A.; van der Voort van Zyp, J. R N|info:eu-repo/dai/nl/326982728; Moerland, M. A.|info:eu-repo/dai/nl/153488174; van Vulpen, M.|info:eu-repo/dai/nl/250581035; Hoekstra, C. J.

    2017-01-01

    Purpose: Determining the independent effect of additional intraoperative adaptive C-arm cone-beam CT (CBCT) planning vs. transrectal ultrasound (TRUS)-guided interactive planning alone in 125I brachytherapy for prostate cancer (PCa) on biochemical disease-free survival (BDFS). Methods and materials:

  7. Validated image fusion of dedicated PET and CT for external beam radiation and therapy in the head and neck area.

    NARCIS (Netherlands)

    Vogel, W.V.; Schinagl, D.A.X.; Dalen, J.A. van; Kaanders, J.H.A.M.; Oyen, W.J.G.

    2008-01-01

    AIM: Integration of positron emission tomography (PET) information into computer tomography (CT)- based intensity modulated external beam radiation therapy (IMRT) allows adaptation of the target volume to functional parameters, but only when the image registration procedure is reliable. The aim of

  8. Influence of object location in different FOVs on trabecular bone microstructure measurements of human mandible: a cone beam CT study

    NARCIS (Netherlands)

    Ibrahim, N.; Parsa, A.; Hassan, B.; van der Stelt, P.; Aartman, I.H.A.; Nambiar, P.

    2014-01-01

    The aim of this study was to assess the influence of different object locations in different fields of view (FOVs) of two cone beam CT (CBCT) systems on trabecular bone microstructure measurements of a human mandible. A block of dry human mandible was scanned at five different locations (centre,

  9. Radiation dose response of normal lung assessed by Cone Beam CT - a potential tool for biologically adaptive radiation therapy

    DEFF Research Database (Denmark)

    Bertelsen, Anders; Schytte, Tine; Bentzen, Søren M

    2011-01-01

    Density changes of healthy lung tissue during radiotherapy as observed by Cone Beam CT (CBCT) might be an early indicator of patient specific lung toxicity. This study investigates the time course of CBCT density changes and tests for a possible correlation with locally delivered dose....

  10. Image and surgery-related costs comparing cone beam CT and panoramic imaging before removal of impacted mandibular third molars

    DEFF Research Database (Denmark)

    Petersen, L. B.; Rose Olsen, Kim; Christensen, J.

    2014-01-01

    Objectives: The aim of this prospective clinical study was to derive the absolute and relative costs of cone beam CT (CBCT) and panoramic imaging before removal of an impacted mandibular third molar. Furthermore, the study aimed to analyse the influence of different cost-setting scenarios...

  11. Combined Fluoroscopy- and CT-Guided Transthoracic Needle Biopsy Using a C-Arm Cone-Beam CT System: Comparison with Fluoroscopy-Guided Biopsy

    Science.gov (United States)

    Cheung, Joo Yeon; Shim, Sung Shine; Lim, Soo Mee

    2011-01-01

    Objective The aim of this study was to evaluate the usefulness of combined fluoroscopy- and CT-guided transthoracic needle biopsy (FC-TNB) using a cone beam CT system in comparison to fluoroscopy-guided TNB (F-TNB). Materials and Methods We retrospectively evaluated 74 FC-TNB cases (group A) and 97 F-TNB cases (group B) to compare their respective diagnostic accuracies according to the size and depth of the lesion, as well as complications, procedure time, and radiation dose. Results The sensitivity for malignancy and diagnostic accuracy for small ( 0.05). Concerning lesions ≥ 30 mm in size and fluoroscopy- and CT-guided TNB allows the biopsy of small (< 30 mm) and deep lesions (≥ 50 mm) with high diagnostic accuracy and short procedure times, whereas F-TNB is still a useful method for large and superficial lesions with a low radiation dose. PMID:21228944

  12. Pulmonary embolism: comprehensive diagnosis by using electron-beam CT for detection of emboli and assessment of pulmonary blood flow.

    Science.gov (United States)

    Schoepf, U J; Bruening, R; Konschitzky, H; Becker, C R; Knez, A; Weber, J; Muehling, O; Herzog, P; Huber, A; Haberl, R; Reiser, M F

    2000-12-01

    To comprehensively assess thoracic anatomy and pulmonary microcirculation in pulmonary embolism by using computed tomographic (CT) angiography of the pulmonary arteries combined with functional CT imaging of blood flow. Twenty-two patients suspected of having acute pulmonary embolism underwent contrast material-enhanced thin-section electron-beam CT angiography of the pulmonary arteries. In addition, in each patient, a dynamic multisection blood flow CT study was performed on a 7.6-cm lung volume with electrocardiographic gating. Pulmonary blood flow was calculated, and perfusion parameters were visualized on color-coded maps. The color-coded maps and CT angiograms were independently evaluated, segment by segment, by two readers for perfusion deficits and the presence of clots, respectively. The results were compared. Mean pulmonary blood flow was 0.63 mL/min/mL in the occluded segments versus 2.27 mL/min/mL in the nonoccluded segments (P: =.001). The sensitivity and specificity of perfusion maps for the presence of segmental pulmonary embolism compared with those of CT angiography were 75.4% and 82.3%, respectively, with positive and negative predictive values of 79.6% and 84.7%, respectively. The false-negative findings were caused mainly by partial occlusion of vessels. In eight patients, a substantial alternative or additional pathologic entity was diagnosed. By combining CT angiography and dynamic CT imaging, a comprehensive and noninvasive diagnosis of thoracic structure and function is feasible with a single modality.

  13. Power cables with extruded insulation and their accessories for rated voltages from 1 kV (Um = 1,2 kV) up to 30 kV (Um = 36 kV) : Part 2: cables for rated voltages from 6 kV (Um = 7,2 kV) up to 30 kV (Um = 36 kV)

    CERN Document Server

    International Electrotechnical Commission. Geneva

    2005-01-01

    Power cables with extruded insulation and their accessories for rated voltages from 1 kV (Um = 1,2 kV) up to 30 kV (Um = 36 kV) : Part 2: cables for rated voltages from 6 kV (Um = 7,2 kV) up to 30 kV (Um = 36 kV)

  14. Diagnostic Accuracy of Split-Bolus Single-Phase Contrast-Enhanced Cone-Beam CT for the Detection of Liver Tumors before Transarterial Chemoembolization.

    Science.gov (United States)

    Jonczyk, Martin; Chapiro, Julius; Collettini, Federico; Geisel, Dominik; Schnapauff, Dirk; Streitparth, Florian; Schmidt, Thomas; Hamm, Bernd; Gebauer, Bernhard; Wieners, Gero

    2017-10-01

    To evaluate detectability of hepatocellular carcinoma (HCC) using split-bolus cone-beam CT in intraindividual comparison between cone-beam CT and contrast-enhanced MR imaging. In a retrospective, single-center study, 28 patients with 85 HCC tumors were treated with transarterial chemoembolization between May 2015 and June 2016. All patients underwent arterial and hepatobiliary phase (HBP) MR imaging within 1 month before transarterial chemoembolization. Cone-beam CT images were acquired using a split-bolus contrast injection with 2 contrast injections and 1 cone-beam CT acquisition. Statistical analyses included Friedman 2-way analysis, Kendall coefficient of concordance, and Wilcoxon test. Tumor detectability was scored using a 5-point system (1 = best; 5 = worst) by 2 independent readers resulting in 170 evaluated tumors. Quantitative analysis included signal-to-noise and contrast-to-noise ratio and contrast measurements. P values cone-beam CT (2.91/2.73) and HBP MR imaging (2.93/2.21) compared with arterial MR imaging (3.72/3.05; P cone-beam CT and HBP MR imaging in terms of detectability (P = .154) and sensitivity for hypervascularized tumors. More tumors were identified on cone-beam CT (n = 121/170) than on arterial MR imaging (n = 94/170). Average contrast-to-noise ratio values of arterial and HBP MR imaging were higher than for cone-beam CT (7.79, 8.58, 4.43), whereas contrast values were higher for cone-beam CT than for MR imaging (0.11, 0.13, 0.97). Split-bolus cone-beam CT showed excellent detectability of HCC. Sensitivity is comparable to HBP MR imaging and better than arterial phase MR imaging. Copyright © 2017 SIR. Published by Elsevier Inc. All rights reserved.

  15. SU-D-BRA-03: Simultaneous MV-KV Imaging for Intra-Fractional Motion Management during Volume Modulated Arc Therapy (VMAT) Delivery on the Varian TrueBeam.

    Science.gov (United States)

    Hunt, M; Regmi, R; Zhang, P; Pham, H; Wang, P; Xiong, J; Morf, D; Mageras, G

    2012-06-01

    To evaluate a MV-kV intra-fractional imaging technique for use during volume modulated arc therapy (VMAT) with the Varian TrueBeam. MV-kV image pairs were acquired intra-fractionally during VMAT delivery. kV images (11 fps) were acquired throughout delivery using a standard pre-programmed imaging template. MV images (9.5 fps) were acquired simultaneously by deploying the EPID and passively collecting the resulting images using Varian proprietary software, iTools Capture. Localization accuracy was evaluated by imaging a Rando phantom implanted with 3 fiducials while moving the couch according to XML- programmed trajectories simulating typical prostate and respiratory motion. VMAT delivery was done using a single 360 degree arc in TrueBeam Developer mode. The effect on accuracy of total MU and gantry speed was studied. To improve image quality, MV frame averaging was performed and the MV and kV images were then registered to their corresponding DRRs using in-house registration software. From these 2D registrations, the 3D position at each MV-kV acquisition point was determined. Between 130 and 390 MV-kV pairs were acquired for each delivery. The mean difference between planned couch and measured fiducial 3D positions with prostate motion was less than 0.03 cm in each direction (SD 0.03 cm). Neither gantry speed nor MU significantly impacted accuracy. for respiratory motion, the mean difference between planned and measured position was less than 0.04 cm. Standard deviation averaged 0.06 cm but increased to 0.12 cm with large instantaneous motion and less MV dose per frame. MV frame averaging and inaccuracies in MV image gantry angle determination also affected accuracy, particularly with significant motion. With high quality MV imaging, MV-kV localization techniques can be highly accurate, even in the presence of significant motion. As clinical MV-kV methods become available, such techniques can provide an efficient and accurate method for monitoring intra

  16. The effects of field-of-view and patient size on CT numbers from cone-beam computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Seet, Katrina Y T; Barghi, Arvand; Yartsev, Slav; Van Dyk, Jake [London Regional Cancer Program, London Health Sciences Centre, London, Ontario (Canada)], E-mail: slav.yartsev@lhsc.on.ca

    2009-10-21

    Cone-beam computed tomography (CBCT) is used for patient alignment before treatment and is ideal for use in adaptive radiotherapy to account for tumor shrinkage, organ deformation and weight loss. However, CBCT images are prone to artifacts such as streaking and cupping effects, reducing image quality and CT number accuracy. Our goal was to determine the optimum combination of cone-beam imaging options to increase the accuracy of image CT numbers. Several phantoms with and without inserts of known relative electron densities were imaged using the Varian on-board imaging system. It was found that CT numbers are most influenced by the selection of field-of-view and are dependent on object size and filter type. Image acquisition in half-fan mode consistently produced more accurate CT numbers, regardless of phantom size. Values measured using full-fan mode can differ by up to 7% from planning CT values. No differences were found between CT numbers of all phantom images with low and standard dose modes.

  17. Perfusion deficit versus anatomic visualization in detection of pulmonary emboli via electron-beam CT: Validation in swine

    Energy Technology Data Exchange (ETDEWEB)

    Hoffman, E.A.; Tajik, J.K.; Petersen, G.; Reiners, T.J.; Thompson, B.H.; Stanford, W. [Univ. of Iowa College of Medicine, Iowa City, IA (United States)

    1995-12-31

    The authors present here the initial findings regarding the utility of functional X-ray CT imaging in determining the presence of pulmonary emboli. Recently, X-ray CT has been reported to be a promising technique in detecting pulmonary emboli through direct visualization of the clot as a filling detect of the reconstructed vascular lumen with CT scanning occurring during i.v. contrast drip. To determine whether functional imaging via the dynamic mode of electron beam CT might add to the sensitivity and specificity of associated temporal parameters, the authors have scanned 32 pigs and report here the findings on 17 pigs evaluated to date. Findings to date show that the evaluation of flow deficits detected via electron beam CT with a small 2--3 sec. bolus contrast injection has the potential to provide improvement in embolus detection over visual inspection of thin section CT/continuous infusion contrast where the viewer is looking for unenhanced regions in the pulmonary arteries.

  18. A One-Step Cone-Beam CT-Enabled Planning-to-Treatment Model for Palliative Radiotherapy-From Development to Implementation

    Energy Technology Data Exchange (ETDEWEB)

    Wong, Rebecca K.S., E-mail: rebecca.wong@rmp.uhn.on.ca [Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Letourneau, Daniel; Varma, Anita [Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Bissonnette, Jean Pierre; Fitzpatrick, David; Grabarz, Daniel; Elder, Christine [Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario (Canada); Martin, Melanie; Bezjak, Andrea [Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Panzarella, Tony [Department of Biostatistics, Princess Margaret Hospital, Toronto, Ontario (Canada); Gospodarowicz, Mary [Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Jaffray, David A. [Radiation Medicine Program, Princess Margaret Hospital, Toronto, Ontario (Canada); Department of Radiation Oncology, University of Toronto, Toronto, Ontario (Canada); Department of Medical Biophysics, University of Toronto, Toronto, Ontario (Canada)

    2012-11-01

    Purpose: To develop a cone-beam computed tomography (CT)-enabled one-step simulation-to-treatment process for the treatment of bone metastases. Methods and Materials: A three-phase prospective study was conducted. Patients requiring palliative radiotherapy to the spine, mediastinum, or abdomen/pelvis suitable for treatment with simple beam geometry ({<=}2 beams) were accrued. Phase A established the accuracy of cone-beam CT images for the purpose of gross tumor target volume (GTV) definition. Phase B evaluated the feasibility of implementing the cone-beam CT-enabled planning process at the treatment unit. Phase C evaluated the online cone-beam CT-enabled process for the planning and treatment of patients requiring radiotherapy for bone metastases. Results: Eighty-four patients participated in this study. Phase A (n = 9) established the adequacy of cone-beam CT images for target definition. Phase B (n = 45) established the quality of treatment plans to be adequate for clinical implementation for bone metastases. When the process was applied clinically in bone metastases (Phase C), the degree of overlap between planning computed tomography (PCT) and cone-beam CT for GTV and between PCT and cone-beam CT for treatment field was 82% {+-} 11% and 97% {+-} 4%, respectively. The oncologist's decision to accept the plan under a time-pressured environment remained of high quality, with the cone-beam CT-generated treatment plan delivering at least 90% of the prescribed dose to 100% {+-} 0% of the cone-beam CT planning target volume (PTV). With the assumption that the PCT PTV is the gold-standard target, the cone-beam CT-generated treatment plan delivered at least 90% and at least 95% of dose to 98% {+-} 2% and 97% {+-} 5% of the PCT PTV, respectively. The mean time for the online planning and treatment process was 32.7 {+-} 4.0 minutes. Patient satisfaction was high, with a trend for superior satisfaction with the cone-beam CT-enabled process. Conclusions: The cone-beam

  19. A one-step cone-beam CT-enabled planning-to-treatment model for palliative radiotherapy-from development to implementation.

    Science.gov (United States)

    Wong, Rebecca K S; Letourneau, Daniel; Varma, Anita; Bissonnette, Jean Pierre; Fitzpatrick, David; Grabarz, Daniel; Elder, Christine; Martin, Melanie; Bezjak, Andrea; Panzarella, Tony; Gospodarowicz, Mary; Jaffray, David A

    2012-11-01

    To develop a cone-beam computed tomography (CT)-enabled one-step simulation-to-treatment process for the treatment of bone metastases. A three-phase prospective study was conducted. Patients requiring palliative radiotherapy to the spine, mediastinum, or abdomen/pelvis suitable for treatment with simple beam geometry (≤2 beams) were accrued. Phase A established the accuracy of cone-beam CT images for the purpose of gross tumor target volume (GTV) definition. Phase B evaluated the feasibility of implementing the cone-beam CT-enabled planning process at the treatment unit. Phase C evaluated the online cone-beam CT-enabled process for the planning and treatment of patients requiring radiotherapy for bone metastases. Eighty-four patients participated in this study. Phase A (n = 9) established the adequacy of cone-beam CT images for target definition. Phase B (n = 45) established the quality of treatment plans to be adequate for clinical implementation for bone metastases. When the process was applied clinically in bone metastases (Phase C), the degree of overlap between planning computed tomography (PCT) and cone-beam CT for GTV and between PCT and cone-beam CT for treatment field was 82% ± 11% and 97% ± 4%, respectively. The oncologist's decision to accept the plan under a time-pressured environment remained of high quality, with the cone-beam CT-generated treatment plan delivering at least 90% of the prescribed dose to 100% ± 0% of the cone-beam CT planning target volume (PTV). With the assumption that the PCT PTV is the gold-standard target, the cone-beam CT-generated treatment plan delivered at least 90% and at least 95% of dose to 98% ± 2% and 97% ± 5% of the PCT PTV, respectively. The mean time for the online planning and treatment process was 32.7 ± 4.0 minutes. Patient satisfaction was high, with a trend for superior satisfaction with the cone-beam CT-enabled process. The cone-beam CT-enabled palliative treatment process is feasible and is ready for

  20. Accuracy of a Cone-Beam CT Virtual Parenchymal Perfusion Algorithm for Liver Cancer Targeting during Intra-arterial Therapy.

    Science.gov (United States)

    Derbel, Haytham; Kobeiter, Hicham; Pizaine, Guillaume; Ridouani, Fourat; Luciani, Alain; Radaelli, Alessandro; Van der Sterren, William; Chiaradia, Mélanie; Tacher, Vania

    2018-02-01

    To evaluate accuracy of virtual parenchymal perfusion (VPP) algorithm developed for targeting liver cancer during intra-arterial therapy (IAT) using cone-beam CT guidance. VPP was retrospectively applied to 15 patients who underwent IAT for liver cancer. Virtual territory (VT) was estimated after positioning a virtual injection point on nonselective dual-phase (DP) cone-beam CT images acquired during hepatic arteriography at the same position chosen for selective treatment. Targeted territory (TT) was used as the gold standard and was defined by parenchymal phase enhancement of selective DP cone-beam CT performed before treatment start. Qualitative evaluation of anatomic conformity between VT and TT was performed using a 3-rank scale (poor, acceptable, excellent) by 3 double-blinded readers. VT and TT were also quantitatively compared using spatial overlap-based (Dice similarity coefficient [DSC], sensitivity, and positive predictive value), distance-based (mean surface distance [MSD]), and volume-based (absolute volume error and correlation between pairwise volumes) metrics. Interreader agreement was evaluated for the 2 evaluation methods. Eighteen DP cone-beam CT scans were performed. Qualitative evaluation showed excellent overlap between VT and TT in 88.9%-94.4%, depending on the readers. DSC was 0.78 ± 0.1, sensitivity was 80%, positive predictive value was 83%, and MSD was 5.1 mm ± 2.4. Absolute volume error was 15%, and R 2 Pearson correlation factor was 0.99. Interreader agreement was good for both qualitative and quantitative evaluations. VPP algorithm is accurate and reliable in identification of liver arterial territories during IAT using cone-beam CT guidance. Copyright © 2017 SIR. Published by Elsevier Inc. All rights reserved.

  1. The European Society of Therapeutic Radiology and Oncology-European Institute of Radiotherapy (ESTRO-EIR) report on 3D CT-based in-room image guidance systems: a practical and technical review and guide

    DEFF Research Database (Denmark)

    Korreman, Stine; Rasch, Coen; McNair, Helen

    2010-01-01

    and technical for treatment delivery and quality assurance. These are reviewed for four categories of solutions; kV CT and kV CBCT (cone-beam CT) as well as MV CT and MV CBCT. It will also provide a framework and checklist to consider the capability and functionality of these systems as well as the resources...... demonstrates a wide variability based on local practices. This report whilst comprehensive is not exhaustive as this area of development remains a very active field for research and development. However, it should serve as a practical guide and framework for all professional groups within the field, focussed...

  2. High-fidelity artifact correction for cone-beam CT imaging of the brain

    Science.gov (United States)

    Sisniega, A.; Zbijewski, W.; Xu, J.; Dang, H.; Stayman, J. W.; Yorkston, J.; Aygun, N.; Koliatsos, V.; Siewerdsen, J. H.

    2015-02-01

    CT is the frontline imaging modality for diagnosis of acute traumatic brain injury (TBI), involving the detection of fresh blood in the brain (contrast of 30-50 HU, detail size down to 1 mm) in a non-contrast-enhanced exam. A dedicated point-of-care imaging system based on cone-beam CT (CBCT) could benefit early detection of TBI and improve direction to appropriate therapy. However, flat-panel detector (FPD) CBCT is challenged by artifacts that degrade contrast resolution and limit application in soft-tissue imaging. We present and evaluate a fairly comprehensive framework for artifact correction to enable soft-tissue brain imaging with FPD CBCT. The framework includes a fast Monte Carlo (MC)-based scatter estimation method complemented by corrections for detector lag, veiling glare, and beam hardening. The fast MC scatter estimation combines GPU acceleration, variance reduction, and simulation with a low number of photon histories and reduced number of projection angles (sparse MC) augmented by kernel de-noising to yield a runtime of ~4 min per scan. Scatter correction is combined with two-pass beam hardening correction. Detector lag correction is based on temporal deconvolution of the measured lag response function. The effects of detector veiling glare are reduced by deconvolution of the glare response function representing the long range tails of the detector point-spread function. The performance of the correction framework is quantified in experiments using a realistic head phantom on a testbench for FPD CBCT. Uncorrected reconstructions were non-diagnostic for soft-tissue imaging tasks in the brain. After processing with the artifact correction framework, image uniformity was substantially improved, and artifacts were reduced to a level that enabled visualization of ~3 mm simulated bleeds throughout the brain. Non-uniformity (cupping) was reduced by a factor of 5, and contrast of simulated bleeds was improved from ~7 to 49.7 HU, in good agreement

  3. Investigation of the accuracy of MV radiation isocentre calculations in the Elekta cone-beam CT software XVI

    DEFF Research Database (Denmark)

    Zimmermann, S. J.; Rowshanfarzad, P.; Ebert, M. A.

    2015-01-01

    Purpose/Objective: Most modern radiotherapy treatments are based on cone-beam CT images to ensure precise positioning of the patient relative to the linac. This requires alignment of the cone-beam CT system to the linac MV radiation isocentre. Therefore, it is important to precisely localize the MV...... and precision of the results. Materials and Methods: A ball bearing phantom with a diameter of 8 mm was attached to the treatment couch positioned close to the linac isocentre. Eight images of the phantom were acquired using the electronic portal imaging device (EPID). Image acquisition was based on the Elekta...... beams. Of the four tested linacs, two were Agility (160 MLC leaves) and two were MLCi2 (80 MLC leaves). The orientation of the MLCs are indicated with IN and CP which refer to collimator angles (-90o, 90o) and (0o,180o), respectively. For MLCi2 and Agility defined fields, the maximum difference...

  4. Metal artefact reduction for a dental cone beam CT image using image segmentation and backprojection filters

    International Nuclear Information System (INIS)

    Mohammadi, Mahdi; Khotanlou, Hassan; Mohammadi, Mohammad

    2011-01-01

    Full text: Due to low dose delivery and fast scanning, the dental Cone Beam CT (CBCT) is the latest technology being implanted for a range of dental imaging. The presence of metallic objects including amalgam or gold fillings in the mouth produces an intuitive image for human jaws. The feasibility of a fast and accurate approach for metal artefact reduction for dental CBCT is investigated. The current study investigates the metal artefact reduction using image segmentation and modification of several sinigrams. In order to reduce metal effects such as beam hardening, streak artefact and intense noises, the application of several algorithms is evaluated. The proposed method includes three stages: preprocessing, reconstruction and post-processing. In the pre-processing stage, in order to reduce the noise level, several phase and frequency filters were applied. At the second stage, based on the specific sinogram achieved for each segment, spline interpolation and weighting backprojection filters were applied to reconstruct the original image. A three-dimensional filter was then applied on reconstructed images, to improve the image quality. Results showed that compared to other available filters, standard frequency filters have a significant influence in the preprocessing stage (ΔHU = 48 ± 6). In addition, with the streak artefact, the probability of beam hardening artefact increases. t e post-processing stage, the application of three-dimensional filters improves the quality of reconstructed images (See Fig. I). Conclusion The proposed method reduces metal artefacts especially where there are more than one metal implanted in the region of interest.

  5. High power semiconductor switches in the 12 kV, 50 kA pulse generator of the SPS beam dump kicker system

    CERN Document Server

    Bonthond, J; Faure, P; Vossenberg, Eugène B

    2001-01-01

    Horizontal deflection of the beam in the dump kicker system of the CERN SPS accelerator is obtained with a series of fast pulsed magnets. The high current pulses of 50 kA per magnet are generated with capacitor discharge type generators which, combined with a resistive free-wheel diode circuit, deliver a critically damped half-sine current with a rise-time of 25 ms. Each generator consists of two 25 kA units, connected in parallel to a magnet via a low inductance transmission line.

  6. Electron-beam lithography simulation for mask making: VI. Comparison of 10- and 50-kV GHOST proximity effect correction

    Science.gov (United States)

    Mack, Chris A.

    2001-09-01

    GHOST uses two exposures, the primary dose and its complement, in an attempt to equalize the effects of backscattering and reduce proximity effects. Unfortunately, image contrast is reduced compared to exposures done without GHOST. A simplified raster scan theory is developed in order to examine the effects of backscattering and GHOST proximity correction on the quality of the images produced. Electron beam lithography simulation is used to examine the effect of spot size and voltage on the spot image generated in 400 nm of ZEP 7000 resist, and the effects of GHOST on proximity effects and process latitude.

  7. Evaluation of deformable image registration for contour propagation between CT and cone-beam CT images in adaptive head and neck radiotherapy.

    Science.gov (United States)

    Li, X; Zhang, Y Y; Shi, Y H; Zhou, L H; Zhen, X

    2016-04-29

    Deformable image registration (DIR) is a critical technic in adaptive radiotherapy (ART) to propagate contours between planning computerized tomography (CT) images and treatment CT/Cone-beam CT (CBCT) image to account for organ deformation for treatment re-planning. To validate the ability and accuracy of DIR algorithms in organ at risk (OAR) contours mapping, seven intensity-based DIR strategies are tested on the planning CT and weekly CBCT images from six Head & Neck cancer patients who underwent a 6 ∼ 7 weeks intensity-modulated radiation therapy (IMRT). Three similarity metrics, i.e. the Dice similarity coefficient (DSC), the percentage error (PE) and the Hausdorff distance (HD), are employed to measure the agreement between the propagated contours and the physician delineated ground truths. It is found that the performance of all the evaluated DIR algorithms declines as the treatment proceeds. No statistically significant performance difference is observed between different DIR algorithms (p> 0.05), except for the double force demons (DFD) which yields the worst result in terms of DSC and PE. For the metric HD, all the DIR algorithms behaved unsatisfactorily with no statistically significant performance difference (p= 0.273). These findings suggested that special care should be taken when utilizing the intensity-based DIR algorithms involved in this study to deform OAR contours between CT and CBCT, especially for those organs with low contrast.

  8. Homogeneous and inhomogeneous material effect in gamma index evaluation of IMRT technique based on fan beam and Cone Beam CT patient images

    Science.gov (United States)

    Wibowo, W. E.; Waliyyulhaq, M.; Pawiro, S. A.

    2017-05-01

    Patient-specific Quality Assurance (QA) technique in lung case Intensity-Modulated Radiation Therapy (IMRT) is traditionally limited to homogeneous material, although the fact that the planning is carried out with inhomogeneous material present. Moreover, the chest area has many of inhomogeneous material, such as lung, soft tissue, and bone, which inhomogeneous material requires special attention to avoid inaccuracies in dose calculation in the Treatment Planning System (TPS). Recent preliminary studies shown that the role of Cone Beam CT (CBCT) can be used not only to position the patient at the time prior to irradiation but also to serve as planning modality. Our study presented the influence of a homogeneous and inhomogeneous materials using Fan Beam CT and Cone Beam CT modalities in IMRT technique on the Gamma Index (GI) value. We used a variation of the segment and Calculation Grid Resolution (CGR). The results showed the deviation of averaged GI value to be between CGR 0.2 cm and 0.4 cm with homogeneous material ranging from -0.44% to 1.46%. For inhomogeneous material, the value was range from -1.74% to 0.98%. In performing patient-specific IMRT QA techniques for lung cancer, homogeneous material can be implemented in evaluating the gamma index.

  9. A new method for x-ray scatter correction: first assessment on a cone-beam CT experimental setup

    Energy Technology Data Exchange (ETDEWEB)

    Rinkel, J [CEA-LETI MINATEC, Division of Micro Technologies for Biology and Healthcare, 38054 Grenoble Cedex 09 (France); Gerfault, L [CEA-LETI MINATEC, Division of Micro Technologies for Biology and Healthcare, 38054 Grenoble Cedex 09 (France); Esteve, F [INSERM U647-RSRM, ESRF, BP200, 38043 Grenoble Cedex 09 (France); Dinten, J-M [CEA-LETI MINATEC, Division of Micro Technologies for Biology and Healthcare, 38054 Grenoble Cedex 09 (France)

    2007-08-07

    Cone-beam computed tomography (CBCT) enables three-dimensional imaging with isotropic resolution and a shorter acquisition time compared to a helical CT scanner. Because a larger object volume is exposed for each projection, scatter levels are much higher than in collimated fan-beam systems, resulting in cupping artifacts, streaks and quantification inaccuracies. In this paper, a general method to correct for scatter in CBCT, without supplementary on-line acquisition, is presented. This method is based on scatter calibration through off-line acquisition combined with on-line analytical transformation based on physical equations, to adapt calibration to the object observed. The method was tested on a PMMA phantom and on an anthropomorphic thorax phantom. The results were validated by comparison to simulation for the PMMA phantom and by comparison to scans obtained on a commercial multi-slice CT scanner for the thorax phantom. Finally, the improvements achieved with the new method were compared to those obtained using a standard beam-stop method. The new method provided results that closely agreed with the simulation and with the conventional CT scanner, eliminating cupping artifacts and significantly improving quantification. Compared to the beam-stop method, lower x-ray doses and shorter acquisition times were needed, both divided by a factor of 9 for the same scatter estimation accuracy.

  10. Fast Data Acquisition in Heavy Ion CT Using Intensifying Screen—EMCCD Camera System With Beam Intensity Monitor

    Science.gov (United States)

    Muraishi, Hiroshi; Abe, Shinji; Satoh, Hitoshi; Hara, Hidetake; Mogaki, Tatsuya; Hara, Satoshi; Miyake, Shoko; Watanabe, Yusuke; Koba, Yusuke

    2012-10-01

    We investigated the feasibility of fast data acquisition in heavy ion CT (IonCT) technique with an X-ray intensifying screen-charged coupled device (CCD) camera system. This technique is based on measuring the residual range distribution of heavy ions after passing through an object. We took a large number of images with a CCD camera for one projection by changing the range shifter (RS) thickness to obtain a characteristic curve similar to a Bragg curve and then to estimate the relative residual range. We used a high quality Electron Multiplying CCD (EMCCD) camera, which drastically reduced data acquisition time. We also used a parallel-plate ionization chamber upstream of an object to monitor the time variation in heavy ion beam intensity from a synchrotron accelerator and to perform beam intensity correction for all EMCCD images. Experiments were conducted using a broad beam of 12C, which was generated by spreading out the pencil beam accelerated up to 400 MeV/u by the Heavy Ion Medical Accelerator, in Chiba (HIMAC) at the National Institute of Radiological Sciences, with a scatterer. We demonstrated that a fast CT data acquisition, 14 min for 256 projections, is possible for an electron density phantom, consisting of six rods with a relative electron density resolution of 0.017, using the proposed technique with HIMAC.

  11. Predicting Cortical Bone Strength from DXA and Dental Cone-Beam CT

    Science.gov (United States)

    Hsu, Jui-Ting; Chen, Ying-Ju; Tsai, Ming-Tzu; Lan, Howard Haw-Chang; Cheng, Fu-Chou; Chen, Michael Y. C.; Wang, Shun-Ping

    2012-01-01

    Objective This study compared the capabilities of dual-energy X-ray absorptiometry (DXA) and dental cone-beam computed tomography (CBCT) for predicting the cortical bone strength of rat femurs and tibias. Materials and Methods Specimens of femurs and tibias obtained from 14 rats were first scanned with DXA to obtain the areal bone mineral density (BMD) of the midshaft cortical portion of the bones. The bones were then scanned using dental CBCT to measure the volumetric cortical bone mineral density (vCtBMD) and the cross-sectional moment of inertia (CSMI) for calculating the bone strength index (BSI). A three-point bending test was conducted to measure the fracture load of each femur and tibia. Bivariate linear Pearson analysis was used to calculate the correlation coefficients (r values) among the CBCT measurements, DXA measurements, and three-point bending parameters. Results The correlation coefficients for the associations of the fracture load with areal BMD (measured using DXA), vCtBMD (measured using CBCT), CSMI (measured using CBCT), and BSI were 0.585 (p = 0.028) and 0.532 (p = 0.050) (for the femur and tibia, respectively), 0.638 (p = 0.014) and 0.762 (p = 0.002), 0.778 (p = 0.001) and 0.792 (pfracture loads in rat femurs and tibias. The BSI, which is a combined index of densitometric and geometric parameters, was especially useful. Further clinical studies are needed to validate the predictive value of BSI obtained from CBCT and should include testing on human cadaver specimens. PMID:23226234

  12. Soft-tissue imaging with C-arm cone-beam CT using statistical reconstruction

    Science.gov (United States)

    Wang, Adam S; Stayman, J Webster; Otake, Yoshito; Kleinszig, Gerhard; Vogt, Sebastian; Gallia, Gary L; Khanna, A Jay; Siewerdsen, Jeffrey H

    2014-01-01

    The potential for statistical image reconstruction methods such as penalized-likelihood (PL) to improve C-arm cone-beam CT (CBCT) soft-tissue visualization for intraoperative imaging over conventional filtered backprojection (FBP) is assessed in this work by making a fair comparison in relation to soft-tissue performance. A prototype mobile C-arm was used to scan anthropomorphic head and abdomen phantoms as well as a cadaveric torso at doses substantially lower than typical values in diagnostic CT, and the effects of dose reduction via tube current reduction and sparse sampling were also compared. Matched spatial resolution between PL and FBP was determined by the edge spread function of low-contrast (~40–80 HU) spheres in the phantoms, which were representative of soft-tissue imaging tasks. PL using the non-quadratic Huber penalty was found to substantially reduce noise relative to FBP, especially at lower spatial resolution where PL provides a contrast-to-noise ratio increase up to 1.4–2.2 × over FBP at 50% dose reduction across all objects. Comparison of sampling strategies indicates that soft-tissue imaging benefits from fully sampled acquisitions at dose above ~1.7 mGy and benefits from 50% sparsity at dose below ~1.0 mGy. Therefore, an appropriate sampling strategy along with the improved low-contrast visualization offered by statistical reconstruction demonstrates the potential for extending intraoperative C-arm CBCT to applications in soft-tissue interventions in neurosurgery as well as thoracic and abdominal surgeries by overcoming conventional tradeoffs in noise, spatial resolution, and dose. PMID:24504126

  13. Clinical implementation of intraoperative cone-beam CT in head and neck surgery

    Science.gov (United States)

    Daly, M. J.; Chan, H.; Nithiananthan, S.; Qiu, J.; Barker, E.; Bachar, G.; Dixon, B. J.; Irish, J. C.; Siewerdsen, J. H.

    2011-03-01

    A prototype mobile C-arm for cone-beam CT (CBCT) has been translated to a prospective clinical trial in head and neck surgery. The flat-panel CBCT C-arm was developed in collaboration with Siemens Healthcare, and demonstrates both sub-mm spatial resolution and soft-tissue visibility at low radiation dose (e.g., <1/5th of a typical diagnostic head CT). CBCT images are available ~15 seconds after scan completion (~1 min acquisition) and reviewed at bedside using custom 3D visualization software based on the open-source Image-Guided Surgery Toolkit (IGSTK). The CBCT C-arm has been successfully deployed in 15 head and neck cases and streamlined into the surgical environment using human factors engineering methods and expert feedback from surgeons, nurses, and anesthetists. Intraoperative imaging is implemented in a manner that maintains operating field sterility, reduces image artifacts (e.g., carbon fiber OR table) and minimizes radiation exposure. Image reviews conducted with surgical staff indicate bony detail and soft-tissue visualization sufficient for intraoperative guidance, with additional artifact management (e.g., metal, scatter) promising further improvements. Clinical trial deployment suggests a role for intraoperative CBCT in guiding complex head and neck surgical tasks, including planning mandible and maxilla resection margins, guiding subcranial and endonasal approaches to skull base tumours, and verifying maxillofacial reconstruction alignment. Ongoing translational research into complimentary image-guidance subsystems include novel methods for real-time tool tracking, fusion of endoscopic video and CBCT, and deformable registration of preoperative volumes and planning contours with intraoperative CBCT.

  14. Adaptive planning using megavoltage fan-beam CT for radiation therapy with testicular shielding

    Energy Technology Data Exchange (ETDEWEB)

    Yadav, Poonam [Department of Human Oncology, University of Wisconsin, Madison, Madison, WI (United States); Department of Medical Physics, University of Wisconsin, Madison, Madison, WI (United States); School of Advance Sciences, Vellore Institue of Technology University, Vellore, Tamil Nadu (India); Kozak, Kevin [Department of Human Oncology, University of Wisconsin, Madison, Madison, WI (United States); Tolakanahalli, Ranjini [Department of Human Oncology, University of Wisconsin, Madison, Madison, WI (United States); Department of Medical Physics, University of Wisconsin, Madison, Madison, WI (United States); Ramasubramanian, V. [School of Advance Sciences, Vellore Institue of Technology University, Vellore, Tamil Nadu (India); Paliwal, Bhudatt R. [Department of Human Oncology, University of Wisconsin, Madison, Madison, WI (United States); Department of Medical Physics, University of Wisconsin, Madison, Madison, WI (United States); University of Wisconsin, Riverview Cancer Centre, Wisconsin Rapids, WI (United States); Welsh, James S. [Department of Human Oncology, University of Wisconsin, Madison, Madison, WI (United States); Department of Medical Physics, University of Wisconsin, Madison, Madison, WI (United States); Rong, Yi, E-mail: rong@humonc.wisc.edu [Department of Human Oncology, University of Wisconsin, Madison, Madison, WI (United States); University of Wisconsin, Riverview Cancer Centre, Wisconsin Rapids, WI (United States)

    2012-07-01

    This study highlights the use of adaptive planning to accommodate testicular shielding in helical tomotherapy for malignancies of the proximal thigh. Two cases of young men with large soft tissue sarcomas of the proximal thigh are presented. After multidisciplinary evaluation, preoperative radiation therapy was recommended. Both patients were referred for sperm banking and lead shields were used to minimize testicular dose during radiation therapy. To minimize imaging artifacts, kilovoltage CT (kVCT) treatment planning was conducted without shielding. Generous hypothetical contours were generated on each 'planning scan' to estimate the location of the lead shield and generate a directionally blocked helical tomotherapy plan. To ensure the accuracy of each plan, megavoltage fan-beam CT (MVCT) scans were obtained at the first treatment and adaptive planning was performed to account for lead shield placement. Two important regions of interest in these cases were femurs and femoral heads. During adaptive planning for the first patient, it was observed that the virtual lead shield contour on kVCT planning images was significantly larger than the actual lead shield used for treatment. However, for the second patient, it was noted that the size of the virtual lead shield contoured on the kVCT image was significantly smaller than the actual shield size. Thus, new adaptive plans based on MVCT images were generated and used for treatment. The planning target volume was underdosed up to 2% and had higher maximum doses without adaptive planning. In conclusion, the treatment of the upper thigh, particularly in young men, presents several clinical challenges, including preservation of gonadal function. In such circumstances, adaptive planning using MVCT can ensure accurate dose delivery even in the presence of high-density testicular shields.

  15. Soft-tissue imaging with C-arm cone-beam CT using statistical reconstruction

    International Nuclear Information System (INIS)

    Wang, Adam S; Stayman, J Webster; Otake, Yoshito; Siewerdsen, Jeffrey H; Kleinszig, Gerhard; Vogt, Sebastian; Gallia, Gary L; Khanna, A Jay

    2014-01-01

    The potential for statistical image reconstruction methods such as penalized-likelihood (PL) to improve C-arm cone-beam CT (CBCT) soft-tissue visualization for intraoperative imaging over conventional filtered backprojection (FBP) is assessed in this work by making a fair comparison in relation to soft-tissue performance. A prototype mobile C-arm was used to scan anthropomorphic head and abdomen phantoms as well as a cadaveric torso at doses substantially lower than typical values in diagnostic CT, and the effects of dose reduction via tube current reduction and sparse sampling were also compared. Matched spatial resolution between PL and FBP was determined by the edge spread function of low-contrast (∼40–80 HU) spheres in the phantoms, which were representative of soft-tissue imaging tasks. PL using the non-quadratic Huber penalty was found to substantially reduce noise relative to FBP, especially at lower spatial resolution where PL provides a contrast-to-noise ratio increase up to 1.4–2.2× over FBP at 50% dose reduction across all objects. Comparison of sampling strategies indicates that soft-tissue imaging benefits from fully sampled acquisitions at dose above ∼1.7 mGy and benefits from 50% sparsity at dose below ∼1.0 mGy. Therefore, an appropriate sampling strategy along with the improved low-contrast visualization offered by statistical reconstruction demonstrates the potential for extending intraoperative C-arm CBCT to applications in soft-tissue interventions in neurosurgery as well as thoracic and abdominal surgeries by overcoming conventional tradeoffs in noise, spatial resolution, and dose. (paper)

  16. A new strategy for online adaptive prostate radiotherapy based on cone-beam CT

    International Nuclear Information System (INIS)

    Boggula, Ramesh; Lorenz, Friedlieb; Lohr, Frank; Wolff, Dirk; Boda-Heggemann, Judit; Hesser, Juergen; Wenz, Frederik; Wertz, Hansjoerg

    2009-01-01

    Interfractional organ motion and patient positioning errors during prostate radiotherapy can have deleterious clinical consequences. It has become clinical practice to re-position the patient with image-guided translational position correction before each treatment to compensate for those errors. However, tilt errors can only be corrected with table corrections in six degrees of freedom or ''full'' adaptive treatment planning strategies. Organ shape deformations can only be corrected by ''full'' plan adaptation. This study evaluates the potential of instant treatment plan adaptation (fast isodose line adaptation with real-time dose manipulating tools) based on cone-beam CT (CBCT) to further improve treatment quality. Using in-house software, CBCTs were modified to approximate a correct density calibration. To evaluate the dosimetric accuracy, dose distributions based on CBCTs were compared with dose distributions calculated on conventional planning CTs (PCT) for four datasets (one inhomogeneous phantom, three patient datasets). To determine the potential dosimetric benefit of a ''full'' plan adaptation over translational position correction, dose distributions were re-optimized using graphical ''online'' dose modification tools for three additional patients' CT-datasets with a substantially distended rectum while the original plans have been created with an empty rectum (single treatment fraction estimates). Absolute dose deviations of up to 51% in comparison to the PCT were observed when uncorrected CBCTs were used for replanning. After density calibration of the CBCTs, 97% of the dose deviations were ≤3% (gamma index: 3%/3 mm). Translational position correction restored the PTV dose (D 95 ) to 73% of the corresponding dose of the reference plan. After plan adaptation, larger improvements of dose restoration to 95% were observed. Additionally, the rectal dose (D 30 ) was further decreased by 42 percentage points (mean of three patient datasets). An accurate dose

  17. Regularization design for high-quality cone-beam CT of intracranial hemorrhage using statistical reconstruction

    Science.gov (United States)

    Dang, H.; Stayman, J. W.; Xu, J.; Sisniega, A.; Zbijewski, W.; Wang, X.; Foos, D. H.; Aygun, N.; Koliatsos, V. E.; Siewerdsen, J. H.

    2016-03-01

    Intracranial hemorrhage (ICH) is associated with pathologies such as hemorrhagic stroke and traumatic brain injury. Multi-detector CT is the current front-line imaging modality for detecting ICH (fresh blood contrast 40-80 HU, down to 1 mm). Flat-panel detector (FPD) cone-beam CT (CBCT) offers a potential alternative with a smaller scanner footprint, greater portability, and lower cost potentially well suited to deployment at the point of care outside standard diagnostic radiology and emergency room settings. Previous studies have suggested reliable detection of ICH down to 3 mm in CBCT using high-fidelity artifact correction and penalized weighted least-squared (PWLS) image reconstruction with a post-artifact-correction noise model. However, ICH reconstructed by traditional image regularization exhibits nonuniform spatial resolution and noise due to interaction between the statistical weights and regularization, which potentially degrades the detectability of ICH. In this work, we propose three regularization methods designed to overcome these challenges. The first two compute spatially varying certainty for uniform spatial resolution and noise, respectively. The third computes spatially varying regularization strength to achieve uniform "detectability," combining both spatial resolution and noise in a manner analogous to a delta-function detection task. Experiments were conducted on a CBCT test-bench, and image quality was evaluated for simulated ICH in different regions of an anthropomorphic head. The first two methods improved the uniformity in spatial resolution and noise compared to traditional regularization. The third exhibited the highest uniformity in detectability among all methods and best overall image quality. The proposed regularization provides a valuable means to achieve uniform image quality in CBCT of ICH and is being incorporated in a CBCT prototype for ICH imaging.

  18. Volume-of-change cone-beam CT for image-guided surgery

    International Nuclear Information System (INIS)

    Lee, Junghoon; Stayman, J Webster; Otake, Yoshito; Schafer, Sebastian; Zbijewski, Wojciech; Khanna, A Jay; Siewerdsen, Jeffrey H; Prince, Jerry L

    2012-01-01

    C-arm cone-beam CT (CBCT) can provide intraoperative 3D imaging capability for surgical guidance, but workflow and radiation dose are the significant barriers to broad utilization. One main reason is that each 3D image acquisition requires a complete scan with a full radiation dose to present a completely new 3D image every time. In this paper, we propose to utilize patient-specific CT or CBCT as prior knowledge to accurately reconstruct the aspects of the region that have changed by the surgical procedure from only a sparse set of x-rays. The proposed methods consist of a 3D–2D registration between the prior volume and a sparse set of intraoperative x-rays, creating digitally reconstructed radiographs (DRRs) from the registered prior volume, computing difference images by subtracting DRRs from the intraoperative x-rays, a penalized likelihood reconstruction of the volume of change (VOC) from the difference images, and finally a fusion of VOC reconstruction with the prior volume to visualize the entire surgical field. When the surgical changes are local and relatively small, the VOC reconstruction involves only a small volume size and a small number of projections, allowing less computation and lower radiation dose than is needed to reconstruct the entire surgical field. We applied this approach to sacroplasty phantom data obtained from a CBCT test bench and vertebroplasty data with a fresh cadaver acquired from a C-arm CBCT system with a flat-panel detector. The VOCs were reconstructed from a varying number of images (10–66 images) and compared to the CBCT ground truth using four different metrics (mean squared error, correlation coefficient, structural similarity index and perceptual difference model). The results show promising reconstruction quality with structural similarity to the ground truth close to 1 even when only 15–20 images were used, allowing dose reduction by the factor of 10–20. (paper)

  19. On-line cone beam CT image guidance for vocal cord tumor targeting

    International Nuclear Information System (INIS)

    Osman, Sarah O.S.; Boer, Hans C.J. de; Astreinidou, Eleftheria; Gangsaas, Anne; Heijmen, Ben J.M.; Levendag, Peter C.

    2009-01-01

    Background and purpose: We are developing a technique for highly focused vocal cord irradiation in early glottic carcinoma to optimally treat a target volume confined to a single cord. This technique, in contrast with the conventional methods, aims at sparing the healthy vocal cord. As such a technique requires sub-mm daily targeting accuracy to be effective, we investigate the accuracy achievable with on-line kV-cone beam CT (CBCT) corrections. Materials and methods: CBCT scans were obtained in 10 early glottic cancer patients in each treatment fraction. The grey value registration available in X-ray volume imaging (XVI) software (Elekta, Synergy) was applied to a volume of interest encompassing the thyroid cartilage. After application of the thus derived corrections, residue displacements with respect to the planning CT scan were measured at clearly identifiable relevant landmarks. The intra- and inter-observer variations were also measured. Results: While before correction the systematic displacements of the vocal cords were as large as 2.4 ± 3.3 mm (cranial-caudal population mean ± SD Σ), daily CBCT registration and correction reduced these values to less than 0.2 ± 0.5 mm in all directions. Random positioning errors (SD σ) were reduced to less than 1 mm. Correcting only for translations and not for rotations did not appreciably affect this accuracy. The residue random displacements partly stem from intra-observer variations (SD = 0.2-0.6 mm). Conclusion: The use of CBCT for daily image guidance in combination with standard mask fixation reduced systematic and random set-up errors of the vocal cords to <1 mm prior to the delivery of each fraction dose. Thus, this facilitates the high targeting precision required for a single vocal cord irradiation.

  20. Consideration of the Effect according to Variation of Material and Respiration in Cone-Beam CT

    International Nuclear Information System (INIS)

    Na, Jun Young; Kim, Jung Mi; Kim, Dae Sup; Kang, Tae Young; Baek, Geum Mun; Kwon, Gyeong Tae

    2012-01-01

    Image Guided Radiation Therapy (IGRT) has been carried out using On-Board Imager system (OBI) in Asan Medical Center. For this reason, This study was to analyze and evaluate the impact on Cone-Beam CT according to variation of material and respiration. This study was to acquire and analyze Cone-Beam CT three times for two material: Cylider acryl (lung equvalent material, diameter 3 cm), Fiducial Marker (using clinic) under Motion Phantom able to adjust respiration pattern randomly was varying period, amplitude and baseline vis-a-vis reference respiration pattern. First, According to a kind of material, when being showed 100% in the acryl and 120% in the Fiducial Marker under the condition of same movement of the motion phantom. Second, According to the respiratory alteration, when being showed 1.13 in the baseline shift 1.8 mm and 1.27 in the baseline shift 3.3 mm for acryl. when being showed 1.01 in 1 sec of period and 1.045 in 2.5 sec of period for acryl. When being showed 0.86 in 0.7 times the standard of amplitude and 1.43 in 1.7 times the standard of amplitude for acryl. when being showed 1.18 in the baseline shift 1.8 mm and 1.34 in the baseline shift 3.3 mm for Fiducial Marker. when being showed 1.0 in 1 sec of period and 1.0 in 2.5 sec of period for Fiducial Marker. When being showed 0.99 in 0.7 times the standard of amplitude and 1.66 in 1.7 times the standard of amplitude for Fiducial Marker. The effect of image size of CBCT was 20% in the case of Fiducial marker. The impact of changes in breathing pattern was minimum 13% - maximum 43% for Arcyl, min. 18% - max. 66% for Fiducial marker. This difference makes serious uncertainty. So, Must be stabilized breathing of patient before acquiring CBCT. also must be monitored breathing of patient in the middle of acquire. If you observe considerable change of breathing when acquiring CBCT. After Image Guided, must be need to check treatment site using fluoroscopy. If a change is too big, re-acquiring CBCT.

  1. Stereotactic radiosurgery for intradural spine tumors using cone-beam CT image guidance.

    Science.gov (United States)

    Monserrate, Andrés; Zussman, Benjamin; Ozpinar, Alp; Niranjan, Ajay; Flickinger, John C; Gerszten, Peter C

    2017-01-01

    OBJECTIVE Cone-beam CT (CBCT) image guidance technology has been widely adopted for spine radiosurgery delivery. There is relatively little experience with spine radiosurgery for intradural tumors using CBCT image guidance. This study prospectively evaluated a series of intradural spine tumors treated with radiosurgery. Patient setup accuracy for spine radiosurgery delivery using CBCT image guidance for intradural spine tumors was determined. METHODS Eighty-two patients with intradural tumors were treated and prospectively evaluated. The positioning deviations of the spine radiosurgery treatments in patients were recorded. Radiosurgery was delivered using a linear accelerator with a beam modulator and CBCT image guidance combined with a robotic couch that allows positioning correction in 3 translational and 3 rotational directions. To measure patient movement, 3 quality assurance CBCTs were performed and recorded in 30 patients: before, halfway, and after the radiosurgery treatment. The positioning data and fused images of planning CT and CBCT from the treatments were analyzed to determine intrafraction patient movements. From each of 3 CBCTs, 3 translational and 3 rotational coordinates were obtained. RESULTS The radiosurgery procedure was successfully completed for all patients. Lesion locations included cervical (22), thoracic (17), lumbar (38), and sacral (5). Tumor histologies included schwannoma (27), neurofibromas (18), meningioma (16), hemangioblastoma (8), and ependymoma (5). The mean prescription dose was 17 Gy (range 12-27 Gy) delivered in 1-3 fractions. At the halfway point of the radiation, the translational variations and standard deviations were 0.4 ± 0.5, 0.5 ± 0.8, and 0.4 ± 0.5 mm in the lateral (x), longitudinal (y), and anteroposterior (z) directions, respectively. Similarly, the variations immediately after treatment were 0.5 ± 0.4, 0.5 ± 0.6, and 0.6 ± 0.5 mm along x, y, and z directions, respectively. The mean rotational angles were 0

  2. Physical performance and image optimization of megavoltage cone-beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Morin, Olivier; Aubry, Jean-Francois; Aubin, Michele; Chen, Josephine; Descovich, Martina; Hashemi, Ali-Bani; Pouliot, Jean [Department of Radiation Oncology, Helen Diller Comprehensive Cancer Center, University of California San Francisco, San Francisco, California 94143 and UCSF/UC Berkeley Joint Graduate Group in Bioengineering, San Francisco, California 94158 (United States); Department of Radiation Oncology, Helen Diller Comprehensive Cancer Center, University of California San Francisco, San Francisco, California 94143 (United States); Siemens Oncology Care Systems, Concord, California 94520 (United States); Department of Radiation Oncology, Helen Diller Comprehensive Cancer Center, University of California San Francisco, San Francisco, California 94143 and UCSF/UC Berkeley Joint Graduate Group in Bioengineering, San Francisco, California 94158 (United States)

    2009-04-15

    Megavoltage cone-beam CT (MVCBCT) is the most recent addition to the in-room CT systems developed for image-guided radiation therapy. The first generation MVCBCT system consists of a 6 MV treatment x-ray beam produced by a conventional linear accelerator equipped with a flat panel amorphous silicon detector. The objective of this study was to evaluate the physical performance of MVCBCT in order to optimize the system acquisition and reconstruction parameters for image quality. MVCBCT acquisitions were performed with the clinical system but images were reconstructed and analyzed with a separate research workstation. The geometrical stability and the positioning accuracy of the system were evaluated by comparing geometrical calibrations routinely performed over a period of 12 months. The beam output and detector intensity stability during MVCBCT acquisition were also evaluated by analyzing in-air acquisitions acquired at different exposure levels. Several system parameters were varied to quantify their impact on image quality including the exposure (2.7, 4.5, 9.0, 18.0, and 54.0 MU), the craniocaudal imaging length (2, 5, 15, and 27.4 cm), the voxel size (0.5, 1, and 2 mm), the slice thickness (1, 3, and 5 mm), and the phantom size. For the reconstruction algorithm, the study investigated the effect of binning, averaging and diffusion filtering of raw projections as well as three different projection filters. A head-sized water cylinder was used to measure and improve the uniformity of MVCBCT images. Inserts of different electron densities were placed in a water cylinder to measure the contrast-to-noise ratio (CNR). The spatial resolution was obtained by measuring the point-spread function of the system using an iterative edge blurring technique. Our results showed that the geometric stability and accuracy of MVCBCT were better than 1 mm over a period of 12 months. Beam intensity variations per projection of up to 35.4% were observed for a 2.7 MU MVCBCT acquisition

  3. System matrix computation vs storage on GPU: A comparative study in cone beam CT.

    Science.gov (United States)

    Matenine, Dmitri; Côté, Geoffroi; Mascolo-Fortin, Julia; Goussard, Yves; Després, Philippe

    2018-02-01

    Iterative reconstruction algorithms in computed tomography (CT) require a fast method for computing the intersection distances between the trajectories of photons and the object, also called ray tracing or system matrix computation. This work focused on the thin-ray model is aimed at comparing different system matrix handling strategies using graphical processing units (GPUs). In this work, the system matrix is modeled by thin rays intersecting a regular grid of box-shaped voxels, known to be an accurate representation of the forward projection operator in CT. However, an uncompressed system matrix exceeds the random access memory (RAM) capacities of typical computers by one order of magnitude or more. Considering the RAM limitations of GPU hardware, several system matrix handling methods were compared: full storage of a compressed system matrix, on-the-fly computation of its coefficients, and partial storage of the system matrix with partial on-the-fly computation. These methods were tested on geometries mimicking a cone beam CT (CBCT) acquisition of a human head. Execution times of three routines of interest were compared: forward projection, backprojection, and ordered-subsets convex (OSC) iteration. A fully stored system matrix yielded the shortest backprojection and OSC iteration times, with a 1.52× acceleration for OSC when compared to the on-the-fly approach. Nevertheless, the maximum problem size was bound by the available GPU RAM and geometrical symmetries. On-the-fly coefficient computation did not require symmetries and was shown to be the fastest for forward projection. It also offered reasonable execution times of about 176.4 ms per view per OSC iteration for a detector of 512 × 448 pixels and a volume of 384 3 voxels, using commodity GPU hardware. Partial system matrix storage has shown a performance similar to the on-the-fly approach, while still relying on symmetries. Partial system matrix storage was shown to yield the lowest relative

  4. A dual modality phantom for cone beam CT and ultrasound image fusion in prostate implant

    International Nuclear Information System (INIS)

    Ng, Angela; Beiki-Ardakan, Akbar; Tong, Shidong; Moseley, Douglas; Siewerdsen, Jeffrey; Jaffray, David; Yeung, Ivan W. T.

    2008-01-01

    In transrectal ultrasound (TRUS) guided prostate seed brachytherapy, TRUS provides good delineation of the prostate while x-ray imaging, e.g., C-arm, gives excellent contrast for seed localization. With the recent availability of cone beam CT (CBCT) technology, the combination of the two imaging modalities may provide an ideal system for intraoperative dosimetric feedback during implantation. A dual modality phantom made of acrylic and copper wire was designed to measure the accuracy and precision of image coregistration between a C-arm based CBCT and 3D TRUS. The phantom was scanned with TRUS and CBCT under the same setup condition. Successive parallel transverse ultrasound (US) images were acquired through manual stepping of the US probe across the phantom at an increment of 1 mm over 7.5 cm. The CBCT imaging was done with three reconstructed slice thicknesses (0.4, 0.8, and 1.6 mm) as well as at three different tilt angles (0 deg., 15 deg., 30 deg. ), and the coregistration between CBCT and US images was done using the Variseed system based on four fiducial markers. Fiducial localization error (FLE), fiducial registration error (FRE), and target registration error (TRE) were calculated for all registered image sets. Results showed that FLE were typically less than 0.4 mm, FRE were less than 0.5 mm, and TRE were typically less than 1 mm within the range of operation for prostate implant (i.e., <6 cm to surface of US probe). An analysis of variance test showed no significant difference in TRE for the CBCT-US fusion among the three slice thicknesses (p=0.37). As a comparison, the experiment was repeated with a US-conventional CT scanner combination. No significant difference in TRE was noted between the US-conventional CT fusion and that for all three CBCT image slice thicknesses (p=0.21). CBCT imaging was also performed at three different C-arm tilt angles of 0 deg., 15 deg., and 30 deg. and reconstructed at a slice thickness of 0.8 mm. There is no significant

  5. A model-based scatter artifacts correction for cone beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Zhao, Wei; Zhu, Jun; Wang, Luyao [Department of Biomedical Engineering, Huazhong University of Science and Technology, Hubei 430074 (China); Vernekohl, Don; Xing, Lei, E-mail: lei@stanford.edu [Department of Radiation Oncology, Stanford University, Stanford, California 94305 (United States)

    2016-04-15

    Purpose: Due to the increased axial coverage of multislice computed tomography (CT) and the introduction of flat detectors, the size of x-ray illumination fields has grown dramatically, causing an increase in scatter radiation. For CT imaging, scatter is a significant issue that introduces shading artifact, streaks, as well as reduced contrast and Hounsfield Units (HU) accuracy. The purpose of this work is to provide a fast and accurate scatter artifacts correction algorithm for cone beam CT (CBCT) imaging. Methods: The method starts with an estimation of coarse scatter profiles for a set of CBCT data in either image domain or projection domain. A denoising algorithm designed specifically for Poisson signals is then applied to derive the final scatter distribution. Qualitative and quantitative evaluations using thorax and abdomen phantoms with Monte Carlo (MC) simulations, experimental Catphan phantom data, and in vivo human data acquired for a clinical image guided radiation therapy were performed. Scatter correction in both projection domain and image domain was conducted and the influences of segmentation method, mismatched attenuation coefficients, and spectrum model as well as parameter selection were also investigated. Results: Results show that the proposed algorithm can significantly reduce scatter artifacts and recover the correct HU in either projection domain or image domain. For the MC thorax phantom study, four-components segmentation yields the best results, while the results of three-components segmentation are still acceptable. The parameters (iteration number K and weight β) affect the accuracy of the scatter correction and the results get improved as K and β increase. It was found that variations in attenuation coefficient accuracies only slightly impact the performance of the proposed processing. For the Catphan phantom data, the mean value over all pixels in the residual image is reduced from −21.8 to −0.2 HU and 0.7 HU for projection

  6. Deformable image registration with local rigidity constraints for cone-beam CT-guided spine surgery

    Science.gov (United States)

    Reaungamornrat, S.; Wang, A. S.; Uneri, A.; Otake, Y.; Khanna, A. J.; Siewerdsen, J. H.

    2014-07-01

    Image-guided spine surgery (IGSS) is associated with reduced co-morbidity and improved surgical outcome. However, precise localization of target anatomy and adjacent nerves and vessels relative to planning information (e.g., device trajectories) can be challenged by anatomical deformation. Rigid registration alone fails to account for deformation associated with changes in spine curvature, and conventional deformable registration fails to account for rigidity of the vertebrae, causing unrealistic distortions in the registered image that can confound high-precision surgery. We developed and evaluated a deformable registration method capable of preserving rigidity of bones while resolving the deformation of surrounding soft tissue. The method aligns preoperative CT to intraoperative cone-beam CT (CBCT) using free-form deformation (FFD) with constraints on rigid body motion imposed according to a simple intensity threshold of bone intensities. The constraints enforced three properties of a rigid transformation—namely, constraints on affinity (AC), orthogonality (OC), and properness (PC). The method also incorporated an injectivity constraint (IC) to preserve topology. Physical experiments involving phantoms, an ovine spine, and a human cadaver as well as digital simulations were performed to evaluate the sensitivity to registration parameters, preservation of rigid body morphology, and overall registration accuracy of constrained FFD in comparison to conventional unconstrained FFD (uFFD) and Demons registration. FFD with orthogonality and injectivity constraints (denoted FFD+OC+IC) demonstrated improved performance compared to uFFD and Demons. Affinity and properness constraints offered little or no additional improvement. The FFD+OC+IC method preserved rigid body morphology at near-ideal values of zero dilatation ({ D} = 0.05, compared to 0.39 and 0.56 for uFFD and Demons, respectively) and shear ({ S} = 0.08, compared to 0.36 and 0.44 for uFFD and Demons

  7. The NOVA project: maximizing beam time efficiency through synergistic analyses of SRμCT data

    Science.gov (United States)

    Schmelzle, Sebastian; Heethoff, Michael; Heuveline, Vincent; Lösel, Philipp; Becker, Jürgen; Beckmann, Felix; Schluenzen, Frank; Hammel, Jörg U.; Kopmann, Andreas; Mexner, Wolfgang; Vogelgesang, Matthias; Jerome, Nicholas Tan; Betz, Oliver; Beutel, Rolf; Wipfler, Benjamin; Blanke, Alexander; Harzsch, Steffen; Hörnig, Marie; Baumbach, Tilo; van de Kamp, Thomas

    2017-09-01

    Beamtime and resulting SRμCT data are a valuable resource for researchers of a broad scientific community in life sciences. Most research groups, however, are only interested in a specific organ and use only a fraction of their data. The rest of the data usually remains untapped. By using a new collaborative approach, the NOVA project (Network for Online Visualization and synergistic Analysis of tomographic data) aims to demonstrate, that more efficient use of the valuable beam time is possible by coordinated research on different organ systems. The biological partners in the project cover different scientific aspects and thus serve as model community for the collaborative approach. As proof of principle, different aspects of insect head morphology will be investigated (e.g., biomechanics of the mouthparts, and neurobiology with the topology of sensory areas). This effort is accomplished by development of advanced analysis tools for the ever-increasing quantity of tomographic datasets. In the preceding project ASTOR, we already successfully demonstrated considerable progress in semi-automatic segmentation and classification of internal structures. Further improvement of these methods is essential for an efficient use of beam time and will be refined in the current NOVAproject. Significant enhancements are also planned at PETRA III beamline p05 to provide all possible contrast modalities in x-ray imaging optimized to biological samples, on the reconstruction algorithms, and the tools for subsequent analyses and management of the data. All improvements made on key technologies within this project will in the long-term be equally beneficial for all users of tomography instrumentations.

  8. Iterative image-domain ring artifact removal in cone-beam CT

    Science.gov (United States)

    Liang, Xiaokun; Zhang, Zhicheng; Niu, Tianye; Yu, Shaode; Wu, Shibin; Li, Zhicheng; Zhang, Huailing; Xie, Yaoqin

    2017-07-01

    Ring artifacts in cone beam computed tomography (CBCT) images are caused by pixel gain variations using flat-panel detectors, and may lead to structured non-uniformities and deterioration of image quality. The purpose of this study is to propose a method of general ring artifact removal in CBCT images. This method is based on the polar coordinate system, where the ring artifacts manifest as stripe artifacts. Using relative total variation, the CBCT images are first smoothed to generate template images with fewer image details and ring artifacts. By subtracting the template images from the CBCT images, residual images with image details and ring artifacts are generated. As the ring artifact manifests as a stripe artifact in a polar coordinate system, the artifact image can be extracted by mean value from the residual image; the image details are generated by subtracting the artifact image from the residual image. Finally, the image details are compensated to the template image to generate the corrected images. The proposed framework is iterated until the differences in the extracted ring artifacts are minimized. We use a 3D Shepp-Logan phantom, Catphan©504 phantom, uniform acrylic cylinder, and images from a head patient to evaluate the proposed method. In the experiments using simulated data, the spatial uniformity is increased by 1.68 times and the structural similarity index is increased from 87.12% to 95.50% using the proposed method. In the experiment using clinical data, our method shows high efficiency in ring artifact removal while preserving the image structure and detail. The iterative approach we propose for ring artifact removal in cone-beam CT is practical and attractive for CBCT guided radiation therapy.

  9. TU-AB-204-00: Advances in Cone-Beam CT and Emerging Applications

    International Nuclear Information System (INIS)

    2015-01-01

    This symposium highlights advanced cone-beam CT (CBCT) technologies in four areas of emerging application in diagnostic imaging and image-guided interventions. Each area includes research that extends the spatial, temporal, and/or contrast resolution characteristics of CBCT beyond conventional limits through advances in scanner technology, acquisition protocols, and 3D image reconstruction techniques. Dr. G. Chen (University of Wisconsin) will present on the topic: Advances in C-arm CBCT for Brain Perfusion Imaging. Stroke is a leading cause of death and disability, and a fraction of people having an acute ischemic stroke are suitable candidates for endovascular therapy. Critical factors that affect both the likelihood of successful revascularization and good clinical outcome are: 1) the time between stroke onset and revascularization; and 2) the ability to distinguish patients who have a small volume of irreversibly injured brain (ischemic core) and a large volume of ischemic but salvageable brain (penumbra) from patients with a large ischemic core and little or no penumbra. Therefore, “time is brain” in the care of the stroke patients. C-arm CBCT systems widely available in angiography suites have the potential to generate non-contrast-enhanced CBCT images to exclude the presence of hemorrhage, time-resolved CBCT angiography to evaluate the site of occlusion and collaterals, and CBCT perfusion parametric images to assess the extent of the ischemic core and penumbra, thereby fulfilling the imaging requirements of a “one-stop-shop” in the angiography suite to reduce the time between onset and revascularization therapy. The challenges and opportunities to advance CBCT technology to fully enable the one-stop-shop C-arm CBCT platform for brain imaging will be discussed. Dr. R. Fahrig (Stanford University) will present on the topic: Advances in C-arm CBCT for Cardiac Interventions. With the goal of providing functional information during cardiac interventions

  10. TU-AB-204-00: Advances in Cone-Beam CT and Emerging Applications

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2015-06-15

    This symposium highlights advanced cone-beam CT (CBCT) technologies in four areas of emerging application in diagnostic imaging and image-guided interventions. Each area includes research that extends the spatial, temporal, and/or contrast resolution characteristics of CBCT beyond conventional limits through advances in scanner technology, acquisition protocols, and 3D image reconstruction techniques. Dr. G. Chen (University of Wisconsin) will present on the topic: Advances in C-arm CBCT for Brain Perfusion Imaging. Stroke is a leading cause of death and disability, and a fraction of people having an acute ischemic stroke are suitable candidates for endovascular therapy. Critical factors that affect both the likelihood of successful revascularization and good clinical outcome are: 1) the time between stroke onset and revascularization; and 2) the ability to distinguish patients who have a small volume of irreversibly injured brain (ischemic core) and a large volume of ischemic but salvageable brain (penumbra) from patients with a large ischemic core and little or no penumbra. Therefore, “time is brain” in the care of the stroke patients. C-arm CBCT systems widely available in angiography suites have the potential to generate non-contrast-enhanced CBCT images to exclude the presence of hemorrhage, time-resolved CBCT angiography to evaluate the site of occlusion and collaterals, and CBCT perfusion parametric images to assess the extent of the ischemic core and penumbra, thereby fulfilling the imaging requirements of a “one-stop-shop” in the angiography suite to reduce the time between onset and revascularization therapy. The challenges and opportunities to advance CBCT technology to fully enable the one-stop-shop C-arm CBCT platform for brain imaging will be discussed. Dr. R. Fahrig (Stanford University) will present on the topic: Advances in C-arm CBCT for Cardiac Interventions. With the goal of providing functional information during cardiac interventions

  11. Dose measurements for dental cone-beam CT: a comparison with MSCT and panoramic imaging

    International Nuclear Information System (INIS)

    Deman, P; Ford, N L; Atwal, P; Duzenli, C; Thakur, Y

    2014-01-01

    To date there is a lack of published information on appropriate methods to determine patient doses from dental cone-beam computed tomography (CBCT) equipment. The goal of this study is to apply and extend the methods recommended in the American Association of Physicists in Medicine (AAPM) Report 111 for CBCT equipment to characterize dose and effective dose for a range of dental imaging equipment. A protocol derived from the one proposed by Dixon et al (2010 Technical Report 111, American Association of Physicist in Medicine, MD, USA), was applied to dose measurements of multi-slice CT, dental CBCT (small and large fields of view (FOV)) and a dental panoramic system. The computed tomography dose index protocol was also performed on the MSCT to compare both methods. The dose distributions in a cylindrical polymethyl methacrylate phantom were characterized using a thimble ionization chamber and Gafchromic™ film (beam profiles). Gafchromic™ films were used to measure the dose distribution in an anthropomorphic phantom. A method was proposed to extend dose estimates to planes superior and inferior to the central plane. The dose normalized to 100 mAs measured in the center of the phantom for the large FOV dental CBCT (11.4 mGy/100 mAs) is two times lower than that of MSCT (20.7 mGy/100 mAs) for the same FOV, but approximately 15 times higher than for a panoramic system (0.6 mGy/100 mAs). The effective dose per scan (in clinical conditions) found for the dental CBCT are 167.60 ± 3.62, 61.30 ± 3.88 and 92.86 ± 7.76 mSv for the Kodak 9000 (fixed scan length of 3.7 cm), and the iCAT Next Generation for 6 cm and 13 cm scan lengths respectively. The method to extend the dose estimates from the central slice to superior and inferior slices indicates a good agreement between theory and measurement. The Gafchromic™ films provided useful beam profile data and 2D distributions of dose in phantom. (paper)

  12. Image quality and dose for a multisource cone-beam CT extremity scanner.

    Science.gov (United States)

    Gang, Grace J; Zbijewski, Wojciech; Mahesh, Mahadevappa; Thawait, Gaurav; Packard, Nathan; Yorkston, John; Demehri, Shadpour; Siewerdsen, Jeffrey H

    2018-01-01

    This work investigates the dose characteristics and image quality of a multisource cone-beam CT scanner dedicated for extremity imaging. The scanner has an x-ray source with three separate anode-cathode units evenly distributed along the longitudinal direction. A nominal scan protocol fires the three sources sequentially, and a total of 600 projections (200 for each source) are acquired over a source-detector orbit of 210 o . Dose was measured using a Farmer chamber in three CTDI phantoms stacked end-to-end. Measurements were performed at the central and four peripheral locations of a CTDI phantom on the axial plane and repeated along the longitudinal direction. The extent of 3D sampling of the three-source configuration was assessed in the Fourier domain through noise power spectrum measurements from air scans and compared with that from a single-source scan. A modified Defrise phantom and anthropomorphic knee and hand phantoms were used for visual assessment of cone-beam artifacts in the reconstructed images. The dose distribution for the three-source configuration exhibits radial asymmetry on the axial plane consistent with a short-scan geometry. Along the longitudinal direction, the highest dose was measured at the central axial plane where the field of view (FOV) from all three sources overlaps and falls off more slowly toward the end compared to a single-source configuration. The extent of 3D sampling is improved throughout the FOV as each source compensates for missing frequencies from the adjacent source. As a result, the reduction in streak and shading artifacts is apparent in the reconstructed images of all three phantoms. The improvement in image quality from the three-source configuration is most pronounced in joint spaces farther from the central axial plane. Initial assessment of the multisource scanner demonstrated the advantages over single-source designs in a compact scanner with large longitudinal FOV. The reduction in cone-beam artifact is

  13. Multimodal registration of three-dimensional maxillodental cone beam CT and photogrammetry data over time.

    Science.gov (United States)

    Bolandzadeh, N; Bischof, W; Flores-Mir, C; Boulanger, P

    2013-01-01

    In recent years, one of the foci of orthodontics has been on systems for the evaluation of treatment results and the tracking of tissue variations over time. This can be accomplished through analysing three-dimensional orthodontic images obtained before and after the treatments. Since complementary information is achieved by integrating multiple imaging modalities, cone beam CT (CBCT) and stereophotogrammetry technologies are used in this study to develop a method for tracking bone, teeth and facial soft-tissue variations over time. We propose a two-phase procedure of multimodal (Phase 1) and multitemporal (Phase 2) registration which aligns images taken from the same patient by different imaging modalities and at different times. Extrinsic (for Phase 1) and intrinsic (for Phase 2) landmark-based registration methods are employed as an initiation for a robust iterative closest points algorithm. Since the mandible moves independently of the upper skull, the registration procedure is applied separately on the mandible and the upper skull. The results show that the signed error distributions of both mandible and skull registrations follow a mixture of two Gaussian distributions, corresponding to alignment errors (due to our method) and temporal change over time. We suggest that the large values among the total registration errors correspond to the temporal change resulting from (1) the effect of treatment (i.e. the orthodontic changes of teeth positions); (2) the biological changes such as teeth growth over time, especially for teenagers; and (3) the segmentation procedure and CBCT precision change over time.

  14. [Evaluation of cone-beam CT in diagnosis of supernumerary teeth in the anterior maxilla].

    Science.gov (United States)

    Wen, Chenni; Li, Guo; Ren, Jiayin; Zheng, Guangning

    2012-08-01

    To evaluate the value of cone-beam CT (CBCT) in the diagnosis and orientation of supernumerary teeth in the anterior maxilla. 195 supernumerary teeth in the anterior maxilla of 146 patients were included, which were examined by CBCT. The number, shape, size, 3-dimensional position, growth direction of the supernumeraries and their relationship with the neighboring teeth were analyzed. The 146 patients aged from 5 to 39, and males were affected more than females in a ratio of 2.95:1. 102 (69.9%) patients had single supernumerary teeth. Of the 195 supernumerary teeth, 126 (64.6%) were near the middle line, 131 (67.2%) were conical, 51 (26.2%) were curved root, 98 (50.3%) were inverted and had a length of (11.97 +/- 2.40) mm. The supernumerary teeth often caused complications. The position of the supernumerary teeth is varied in the maxilla, and often causes permanent dentition complications. CBCT imaging yields accurate 3-dimensional pictures of supernumerary teeth, local dental and bony structures, which is helpful for diagnosis and orientation of supernumerary teeth.

  15. Virtual colonoscopy with electron beam CT: correlation with barium enema, colonoscopy and pathology

    International Nuclear Information System (INIS)

    Hong, Hye Suk; Kim, Min Jung; Chung, Jae Joon; Kim, Myeong Jin; Lee, Jong Tae; Yoo, Hyung Sik

    1998-01-01

    To perform virtual colonoscopy using electron beam tomography(EBT) in patients in whom a colonic mass was present, and to compare the results with those obtained using barium enema, colonoscopy and gross pathologic specimens. Materials and Methods : Ten patients in whom colonic masses were diagnosed by either barium enema or colonoscopy were involved in this study. There were nine cases of adenocarcinoma and one of tubulovillous adenoma. Using EBT preoperative abdominopelvic CT scans were performed. Axial scans were then three-dimensionally reconstructed to produce virtual colonoscopic images and were compared with barium enema, colonoscopy and gross pathologic specimens. Virtual colonoscopic images of the masses were classified as either 1)polyploid, 2)sessile,3)fungating, or 4)annular constrictive. We also determined whether ulcers were present within the lesions and whether there was obstruction. Results : After virtual colonoscopy, two lesions were classified as polyploid, one as sessile, five as fungating and two as annular constrictive. Virtual colonoscopic images showed good correlation with the findings of barium enema, colonoscopy and gross pathologic specimens. Three of six ulcerative lesions were observed on colonoscopy; in seven adenocarcinomas with partial or total luminal obstruction, virtual colonoscopy visualized the colon beyond the obstructed sites. In one case, barium contrast failed to pass through the obstructed portion and in six cases, the colonoscope similarly failed. Conclusion : Virtual colonoscopies correlated well with barium enema, colonoscopy and gross pathologic specimens. They provide three dimensional images of colonic masses and are helpful for the evaluation of obstructive lesions

  16. Prostate image-guided radiotherapy by megavolt cone-beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Zucca, Sergio; Carau, Barbara; Solla, Ignazio; Garibaldi, Elisabetta; Farace, Paolo; Lay, Giancarlo; Meleddu, Gianfranco; Gabriele, Pietro [Regional Oncological Hospital, Cagliari (Italy). Dept. of Radiooncology

    2011-08-15

    To test megavolt cone-beam CT (MV-CBCT) in order to evaluate setup errors in prostate radiotherapy. The setup of 9 patients was verified weekly by electronic portal imaging (EPI) and MV-CBCT, both performed in the same treatment session. EPI were compared with digitally reconstructed radiographies (DRRs). MV-CBCTs were matched to simulation CTs by manual registration based on bone markers (BMR), by manual registration based on soft tissues (STR) - rectum, bladder, and seminal vesicles - and by automatic registration (AR) performed by a mutual information algorithm. Shifts were evaluated along the three main axes: anteroposterior (AP), craniocaudal (CC), and laterolateral (LL). Finally, in 4 additional patients showing intraprostatic calcifications, the calcification mismatch error was used to evaluate the three MV-CBCT matching methods. A total of 50 pairs of orthogonal EPIs and 50 MV-CBCTs were analyzed. Assuming an overall tolerance of 2 mm, no significant differences were observed comparing EPI vs BMR in any axis. A significant difference (p < 0.001) was observed along the AP axis comparing EPI vs AR and EPI vs STR. On the calcification data set (22 measures), the calcification mismatch along the AP direction was significantly lower (p < 0.05) after STR than after BMR or AR. Bone markers were not an effective surrogate of the target position and significant differences were observed comparing EPI or BMR vs STR, supporting the assessment of soft tissue position by MVCBs to verify and correct patient setup in prostate radiotherapy. (orig.)

  17. Prostate image-guided radiotherapy by megavolt cone-beam CT

    International Nuclear Information System (INIS)

    Zucca, Sergio; Carau, Barbara; Solla, Ignazio; Garibaldi, Elisabetta; Farace, Paolo; Lay, Giancarlo; Meleddu, Gianfranco; Gabriele, Pietro

    2011-01-01

    To test megavolt cone-beam CT (MV-CBCT) in order to evaluate setup errors in prostate radiotherapy. The setup of 9 patients was verified weekly by electronic portal imaging (EPI) and MV-CBCT, both performed in the same treatment session. EPI were compared with digitally reconstructed radiographies (DRRs). MV-CBCTs were matched to simulation CTs by manual registration based on bone markers (BMR), by manual registration based on soft tissues (STR) - rectum, bladder, and seminal vesicles - and by automatic registration (AR) performed by a mutual information algorithm. Shifts were evaluated along the three main axes: anteroposterior (AP), craniocaudal (CC), and laterolateral (LL). Finally, in 4 additional patients showing intraprostatic calcifications, the calcification mismatch error was used to evaluate the three MV-CBCT matching methods. A total of 50 pairs of orthogonal EPIs and 50 MV-CBCTs were analyzed. Assuming an overall tolerance of 2 mm, no significant differences were observed comparing EPI vs BMR in any axis. A significant difference (p < 0.001) was observed along the AP axis comparing EPI vs AR and EPI vs STR. On the calcification data set (22 measures), the calcification mismatch along the AP direction was significantly lower (p < 0.05) after STR than after BMR or AR. Bone markers were not an effective surrogate of the target position and significant differences were observed comparing EPI or BMR vs STR, supporting the assessment of soft tissue position by MVCBs to verify and correct patient setup in prostate radiotherapy. (orig.)

  18. Three-dimensional focus of attention for iterative cone-beam micro-CT reconstruction

    International Nuclear Information System (INIS)

    Benson, T M; Gregor, J

    2006-01-01

    Three-dimensional iterative reconstruction of high-resolution, circular orbit cone-beam x-ray CT data is often considered impractical due to the demand for vast amounts of computer cycles and associated memory. In this paper, we show that the computational burden can be reduced by limiting the reconstruction to a small, well-defined portion of the image volume. We first discuss using the support region defined by the set of voxels covered by all of the projection views. We then present a data-driven preprocessing technique called focus of attention that heuristically separates both image and projection data into object and background before reconstruction, thereby further reducing the reconstruction region of interest. We present experimental results for both methods based on mouse data and a parallelized implementation of the SIRT algorithm. The computational savings associated with the support region are substantial. However, the results for focus of attention are even more impressive in that only about one quarter of the computer cycles and memory are needed compared with reconstruction of the entire image volume. The image quality is not compromised by either method

  19. The assessment of impacted maxillary canine position with panoramic radiography and cone beam CT.

    Science.gov (United States)

    Jung, Y H; Liang, H; Benson, B W; Flint, D J; Cho, B H

    2012-07-01

    The aim of this study was to correlate the position of impacted maxillary canines on panoramic radiography with cone beam CT (CBCT) and analyse the labiopalatal position of canines and root resorption of permanent incisors in CBCT according to the mesiodistal position of canines on panoramic radiographs. This study was a retrospective radiographic review of 63 patients with 73 impacted maxillary canines. The mesiodistal position of the canine cusp tip was classified by sector location and analysed on 73 impacted canines from 63 panoramic radiographs. The labiopalatal position of the impacted canines and root resorption of permanent incisors were evaluated with CBCT. The sector location on panoramic radiographs was compared with the labiopalatal position of impacted maxillary canines on CBCT. The statistical correlation between panoramic and CBCT findings was examined using the χ(2) test and the Fisher's exact test. Labially impacted canines in CBCT were more frequent in Panoramic Sectors 1, 2 and 3, mid-alveolus impacted canines were more frequent in Sector 4 and palatally impacted canines were more frequent in Sector 5. There was a statistically significant association between the panoramic sectors of the impacted canines and the labiopalatal position of the canines (p panoramic radiography.

  20. [Effectiveness assessment of 3-D cone beam CT used in human bite marks identification].

    Science.gov (United States)

    Wu, Yan; Chen, Xinmin; Shen, Yun; Yu, Jinhao; Tang, Ying; Zhang, Yiming; Zhu, Lei; Xu, Yuanzhi

    2013-02-01

    The present study was aimed to use the 3-D cone beam CT (CBCT) as a new method in human bite marks identification which was carried out in experimental pigskin to assess its effectiveness in our laboratory. Bite marks were digital photographed according to American Board of Forensic Odontology (ABFO) guidelines. In this study, the data of the suspect's dental casts were collected by scanning in two ways: one was after plate scanning, in which the comparison overlays were generated by Adobe Photoshop 8.0 software; the other was by CBCT, which generated comparison overlays automatically. The bite marks were blind identified with the two kinds of data of the suspect's dental casts respectively. ROC curve was used to analyze the sensitivity, specificity, and 95% confidence interval. The results showed that CBCT method got a larger area under the ROC curve: 0.784 (SE = 0.074, 95% CI = 0.639-0.929), and got a very high specificity (specificity 98.7%, 95% CI = 94.5%-99.8%). Thus, this study illustrates that the CBCT used in bite mark identification is an effective and accurate tool and has stronger ability to exclude suspects compared with the conventional method, but the comparison process needs further study to enhance its effectiveness in bite mark identification.

  1. The effect of cone beam CT (CBCT) on therapeutic decision-making in endodontics.

    Science.gov (United States)

    Mota de Almeida, F J; Knutsson, K; Flygare, L

    2014-01-01

    The aim was to assess to what extent cone beam CT (CBCT) used in accordance with current European Commission guidelines in a normal clinical setting has an impact on therapeutic decisions in a population referred for endodontic problems. The study includes data of consecutively examined patients collected from October 2011 to December 2012. From 2 different endodontic specialist clinics, 57 patients were referred for a CBCT examination using criteria in accordance with current European guidelines. The CBCT examinations were performed using similar equipment and standardized among clinics. After a thorough clinical examination, but before CBCT, the examiner made a preliminary therapy plan which was recorded. After the CBCT examination, the same examiner made a new therapy plan. Therapy plans both before and after the CBCT examination were plotted for 53 patients and 81 teeth. As four patients had incomplete protocols, they were not included in the final analysis. 4% of the patients referred to endodontic clinics during the study period were examined with CBCT. The most frequent reason for referral to CBCT examination was to differentiate pathology from normal anatomy, this was the case in 24 patients (45% of the cases). The primary outcome was therapy plan changes that could be attributed to CBCT examination. There were changes in 28 patients (53%). CBCT has a significant impact on therapeutic decision efficacy in endodontics when used in concordance with the current European Commission guidelines.

  2. Electron beam CT diagnosis of congenital unilateral absence of pulmonary artery

    International Nuclear Information System (INIS)

    Zhou Yuan; Dai Ruping; Cao Cheng; Zhang Gejun; Jing Baolian

    2003-01-01

    Objective: To evaluate the clinical value of electron beam CT (EBCT) in diagnosing congenital unilateral absence of pulmonary artery (UAPA). Methods: Patients with clinically suspected pulmonary artery disease or primary pulmonary hypertension underwent EBCT scanning. EBCT confirmed the diagnosis of UAPA in 11 patients, who were also evaluated with echocardiography and chest roentgenography. Cardioangiography and nuclear ventilation-perfusion scan were performed in some patients for a comparative study. Results: 4 female adults had UAPA with out associated congenital anomaly. 3 male children with coexisting complex congenital abnormality had unilateral absence of the left pulmonary artery and 4 patients coexisted other simple cardiovascular anomaly. EBCT scanning simultaneously displayed topographic pattern of both unilateral absence of pulmonary artery and coexisting congenital cardiovascular anomaly, as well as lung diseases. Conclusion: UAPA diagnosed in childhood usually has unilateral absence of the left pulmonary artery and associated congenital cardiovascular anomaly, while UAPA diagnosed in adult usually has UAPA on the right side without associated congenital anomaly. EBCT is one of the optimal imaging techniques in diagnosing UAPA and it greatly increases the diagnostic efficacy than echocardiography dose. Both EBCT and cardioangiography have their own advantages, however, EBCT, as a noninvasive method, should be complementary and not exclusive

  3. Clinical Implementation Of Megavoltage Cone Beam CT As Part Of An IGRT Program

    International Nuclear Information System (INIS)

    Gonzalez, Albin; Kinney, Vicki; Crooks, Cheryl; Bauer, Lisa

    2008-01-01

    Knowing where the tumor is at all times during treatment is the next challenge in the field of radiation therapy. This issue has become more important because with treatments such as Intensity Modulated Radiation Therapy (IMRT), healthy tissue is spared by using very tight margins around the tumor. These tight margins leave very small room for patient setup errors. The use of an imaging modality in the treatment room as a way to localize the tumor for patient set up is generally known as ''Image Guided Radiation Therapy'' or IGRT. This article deals with a form of IGRT known as Megavoltage Cone Beam Computed Tomography (MCBCT) using a Siemens Oncor linear accelerator currently in use at Firelands Regional Medical Center. With MCBCT, we are capable of acquiring CT images right before the treatment of the patient and then use this information to position the patient tumor according to the treatment plan. This article presents the steps followed in order to clinically implement this system, as well as some of the quality assurance tests suggested by the manufacturer and some tests developed in house

  4. Predicting factors for conversion from fluoroscopy guided Percutaneous transthoracic needle biopsy to cone-beam CT guided Percutaneous transthoracic needle biopsy

    International Nuclear Information System (INIS)

    Lee, Kang Ji; Han, Young Min; Jin, Gong Yong; Song, Ji Soo

    2015-01-01

    To evaluate the predicting factors for conversion from fluoroscopy guided percutaneous transthoracic needle biopsy (PTNB) to cone-beam CT guided PTNB. From January 2011 to December 2012, we retrospectively identified 38 patients who underwent cone-beam CT guided PTNB with solid pulmonary lesions, and 76 patients who underwent fluoroscopy guided PTNB were matched to the patients who underwent cone-beam CT guided PTNB for age, sex, and lesion location. We evaluated predicting factors such as, long-axis diameter, short-axis diameter, anterior-posterior diameter, and CT attenuation value of the solid pulmonary lesion affecting conversion from fluoroscopy guided PTNB to cone-beam CT guided PTNB. Pearson χ 2 test, Fisher exact test, and independent t test were used in statistical analyses; in addition, we also used receiver operating characteristics curve to find the proper cut-off values affecting the conversion to cone-beam CT guided PTNB. Short-axis, long-axis, anterior-posterior diameter and CT attenuation value of the solid pulmonary lesion in patients who underwent fluoroscopy guided PTNB were 2.70 ± 1.57 cm, 3.40 ± 1.92 cm, 3.06 ± 1.81 cm, and 35.67 ± 15.70 Hounsfield unit (HU), respectively. Short-axis, long-axis, anterior-posterior diameter and CT attenuation value of the solid pulmonary lesion in patients who underwent cone-beam CT guided PTNB were 1.60 ± 1.30 cm, 2.20 ± 1.45 cm, 1.91 ± 1.99 cm, and 18.32 ± 23.11 HU, respectively. Short-axis, long-axis, anterior-posterior diameter, and CT attenuation value showed a significantly different mean value between the 2 groups (p = 0.001, p < 0.001, p = 0.003, p < 0.001, respectively). Odd ratios of CT attenuation value and short-axis diameter of the solid pulmonary lesion were 0.952 and 0.618, respectively. Proper cut-off values affecting the conversion to cone-beam CT guided PTNB were 1.65 cm (sensitivity 68.4%, specificity 71.1%) in short-axis diameter and 29.50 HU (sensitivity 65.8%, specificity 65.8%) in

  5. SU-F-I-06: Evaluation of Imaging Dose for Modulation Layer Based Dual Energy Cone-Beam CT

    International Nuclear Information System (INIS)

    Ju, Eunbin; Ahn, SoHyun; Cho, Samju; Keum, Ki Chang; Lee, Rena

    2016-01-01

    Purpose: Dual energy cone beam CT system is finding a variety of promising applications in diagnostic CT, both in imaging of endogenous materials and exogenous materials across a range of body sites. Dual energy cone beam CT system to suggest in this study acquire image by rotating 360 degree with half of the X-ray window covered using copper modulation layer. In the region that covered by modulation layer absorb the low energy X-ray by modulation layer. Relative high energy X-ray passes through the layer and contributes to image reconstruction. Dose evaluation should be carried out in order to utilize such an imaging acquirement technology for clinical use. Methods: For evaluating imaging dose of modulation layer based dual energy cone beam CT system, Prototype cone beam CT that configured X-ray tube (D054SB, Toshiba, Japan) and detector (PaxScan 2520V, Varian Medical Systems, Palo Alto, CA) is used. A range of 0.5–2.0 mm thickness of modulation layer is implemented in Monte Carlo simulation (MCNPX, ver. 2.6.0, Los Alamos National Laboratory, USA) with half of X-ray window covered. In-house phantom using in this study that has 3 cylindrical phantoms configured water, Teflon air with PMMA covered for verifying the comparability the various material in human body and is implemented in Monte Carlo simulation. The actual dose with 2.0 mm copper covered half of X-ray window is measured using Gafchromic EBT3 film with 5.0 mm bolus for compared with simulative dose. Results: Dose in phantom reduced 33% by copper modulation layer of 2.0 mm. Scattering dose occurred in modulation layer by Compton scattering effect is 0.04% of overall dose. Conclusion: Modulation layer of that based dual energy cone beam CT has not influence on unnecessary scatter dose. This study was supported by the Radiation Safety Research Programs (1305033) through the Nuclear Safety and Security Commission.

  6. Upper airway alterations/abnormalities in a case series of obstructive sleep apnea patients identified with cone-beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Shigeta, Y.; Shintaku, W.H.; Clark, G.T. [Orofacial Pain/Oral Medicine Center, Div. of Diagnostic Sciences, School of Dentistry, Univ. of Southern California, Los Angeles, CA (United States); Enciso, R. [Div. of Craniofacial Sciences and Therapeutics, School of Dentistry, Univ. of Southern California, Los Angeles, CA (United States); Ogawa, T. [Dept. of Fixed Prosthodontic Dentistry, Tsurumi Univ., School of Dental Medicine, Tsurumi (Japan)

    2007-06-15

    There are many factors that influence the configuration of the upper airway and may contribute to the development of obstructive sleep apnea (OSA). This paper presents a series of 12 consecutive OSA cases where various upper airway alteration/abnormalities were identified using 3D anatomic reconstructions generated from cone-beam CT (CBCT) images. Some cases exhibited more than one type of abnormality and below we describe each of the six types identified with CBCT in this case series. (orig.)

  7. SU-D-206-06: Task-Specific Optimization of Scintillator Thickness for CMOS-Detector Based Cone-Beam Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Vedantham, S; Shrestha, S; Shi, L; Vijayaraghavan, G; Karellas, A [University of Massachusetts Medical School, Worcester, MA (United States)

    2016-06-15

    Purpose: To optimize the cesium iodide (CsI:Tl) scintillator thickness in a complimentary metal-oxide semiconductor (CMOS)-based detector for use in dedicated cone-beam breast CT. Methods: The imaging task considered was the detection of a microcalcification cluster comprising six 220µm diameter calcium carbonate spheres, arranged in the form of a regular pentagon with 2 mm spacing on its sides and a central calcification, similar to that in ACR-recommended mammography accreditation phantom, at a mean glandular dose of 4.5 mGy. Generalized parallel-cascades based linear systems analysis was used to determine Fourier-domain image quality metrics in reconstructed object space, from which the detectability index inclusive of anatomical noise was determined for a non-prewhitening numerical observer. For 300 projections over 2π, magnification-associated focal-spot blur, Monte Carlo derived x-ray scatter, K-fluorescent emission and reabsorption within CsI:Tl, CsI:Tl quantum efficiency and optical blur, fiberoptic plate transmission efficiency and blur, CMOS quantum efficiency, pixel aperture function and additive noise, and filtered back-projection to isotropic 105µm voxel pitch with bilinear interpolation were modeled. Imaging geometry of a clinical prototype breast CT system, a 60 kV Cu/Al filtered x-ray spectrum from 0.3 mm focal spot incident on a 14 cm diameter semi-ellipsoidal breast were used to determine the detectability index for 300–600 µm thick (75µm increments) CsI:Tl. The CsI:Tl thickness that maximized the detectability index was considered optimal. Results: The limiting resolution (10% modulation transfer function, MTF) progressively decreased with increasing CsI:Tl thickness. The zero-frequency detective quantum efficiency, DQE(0), in projection space increased with increasing CsI:Tl thickness. The maximum detectability index was achieved with 525µm thick CsI:Tl scintillator. Reduced MTF at mid-to-high frequencies for 600µm thick CsI:Tl lowered

  8. Cone Beam CT Imaging Analysis of Interfractional Variations in Bladder Volume and Position During Radiotherapy for Bladder Cancer

    International Nuclear Information System (INIS)

    Yee, Don; Parliament, Matthew; Rathee, Satyapal; Ghosh, Sunita; Ko, Lawrence; Murray, Brad

    2010-01-01

    Purpose: To quantify daily bladder size and position variations during bladder cancer radiotherapy. Methods and Materials: Ten bladder cancer patients underwent daily cone beam CT (CBCT) imaging of the bladder during radiotherapy. Bladder and planning target volumes (bladder/PTV) from CBCT and planning CT scans were compared with respect to bladder center-of-mass shifts in the x (lateral), y (anterior-posterior), and z (superior-inferior) coordinates, bladder/PTV size, bladder/PTV margin positions, overlapping areas, and mutually exclusive regions. Results: A total of 262 CBCT images were obtained from 10 bladder cancer patients. Bladder center of mass shifted most in the y coordinate (mean, -0.32 cm). The anterior bladder wall shifted the most (mean, -0.58 cm). Mean ratios of CBCT-derived bladder and PTV volumes to planning CT-derived counterparts were 0.83 and 0.88. The mean CBCT-derived bladder volume (± standard deviation [SD]) outside the planning CT counterpart was 29.24 cm 3 (SD, 29.71 cm 3 ). The mean planning CT-derived bladder volume outside the CBCT counterpart was 47.74 cm 3 (SD, 21.64 cm 3 ). The mean CBCT PTV outside the planning CT-derived PTV was 47.35 cm 3 (SD, 36.51 cm 3 ). The mean planning CT-derived PTV outside the CBCT-derived PTV was 93.16 cm 3 (SD, 50.21). The mean CBCT-derived bladder volume outside the planning PTV was 2.41 cm 3 (SD, 3.97 cm 3 ). CBCT bladder/ PTV volumes significantly differed from planning CT counterparts (p = 0.047). Conclusions: Significant variations in bladder and PTV volume and position occurred in patients in this trial.

  9. Use of cone-beam CT and live 3-D needle guidance to facilitate percutaneous nephrostomy and nephrolithotripsy access in children and adolescents

    Energy Technology Data Exchange (ETDEWEB)

    Hawkins, C.M. [Emory University School of Medicine, Department of Radiology and Imaging Sciences, Atlanta, GA (United States); Kukreja, Kamlesh [Texas Children' s Hospital, Department of Radiology, Houston, TX (United States); Singewald, Timothy; Johnson, Neil D.; Racadio, John M. [Cincinnati Children' s Hospital Medical Center, Department of Radiology, Cincinnati, OH (United States); Minevich, Eugene; Reddy, Pramod [Cincinnati Children' s Hospital Medical Center, Department of Urology, Cincinnati, OH (United States)

    2016-04-15

    Gaining access into non-dilated renal collecting systems for percutaneous nephrolithotripsy, particularly in patients with prohibitive body habitus and/or scoliosis, is often challenging using conventional techniques. To evaluate the feasibility of cone-beam CT for percutaneous nephrostomy placement for subsequent percutaneous nephrolithotripsy in children and adolescents. A retrospective review of percutaneous nephrostomy revealed use of cone-beam CT and 3-D guidance in 12 percutaneous nephrostomy procedures for 9 patients between 2006 and 2015. All cone-beam CT-guided percutaneous nephrostomies were for pre-lithotripsy access and all 12 were placed in non-dilated collecting systems. Technical success was 100%. There were no complications. Cone-beam CT with 3-D guidance is a technically feasible technique for percutaneous nephrostomy in children and adolescents, specifically for nephrolithotripsy access in non-dilated collecting systems. (orig.)

  10. Use of cone-beam CT and live 3-D needle guidance to facilitate percutaneous nephrostomy and nephrolithotripsy access in children and adolescents.

    Science.gov (United States)

    Hawkins, C Matthew; Kukreja, Kamlesh; Singewald, Timothy; Minevich, Eugene; Johnson, Neil D; Reddy, Pramod; Racadio, John M

    2016-04-01

    Gaining access into non-dilated renal collecting systems for percutaneous nephrolithotripsy, particularly in patients with prohibitive body habitus and/or scoliosis, is often challenging using conventional techniques. To evaluate the feasibility of cone-beam CT for percutaneous nephrostomy placement for subsequent percutaneous nephrolithotripsy in children and adolescents. A retrospective review of percutaneous nephrostomy revealed use of cone-beam CT and 3-D guidance in 12 percutaneous nephrostomy procedures for 9 patients between 2006 and 2015. All cone-beam CT-guided percutaneous nephrostomies were for pre-lithotripsy access and all 12 were placed in non-dilated collecting systems. Technical success was 100%. There were no complications. Cone-beam CT with 3-D guidance is a technically feasible technique for percutaneous nephrostomy in children and adolescents, specifically for nephrolithotripsy access in non-dilated collecting systems.

  11. Analytic image reconstruction from partial data for a single-scan cone-beam CT with scatter correction

    Energy Technology Data Exchange (ETDEWEB)

    Min, Jonghwan; Pua, Rizza; Cho, Seungryong, E-mail: scho@kaist.ac.kr [Department of Nuclear and Quantum Engineering, Korea Advanced Institute of Science and Technology (KAIST), 291 Daehak-ro, Yuseong-gu, Daejeon 305-701 (Korea, Republic of); Kim, Insoo; Han, Bumsoo [EB Tech, Co., Ltd., 550 Yongsan-dong, Yuseong-gu, Daejeon 305-500 (Korea, Republic of)

    2015-11-15

    Purpose: A beam-blocker composed of multiple strips is a useful gadget for scatter correction and/or for dose reduction in cone-beam CT (CBCT). However, the use of such a beam-blocker would yield cone-beam data that can be challenging for accurate image reconstruction from a single scan in the filtered-backprojection framework. The focus of the work was to develop an analytic image reconstruction method for CBCT that can be directly applied to partially blocked cone-beam data in conjunction with the scatter correction. Methods: The authors developed a rebinned backprojection-filteration (BPF) algorithm for reconstructing images from the partially blocked cone-beam data in a circular scan. The authors also proposed a beam-blocking geometry considering data redundancy such that an efficient scatter estimate can be acquired and sufficient data for BPF image reconstruction can be secured at the same time from a single scan without using any blocker motion. Additionally, scatter correction method and noise reduction scheme have been developed. The authors have performed both simulation and experimental studies to validate the rebinned BPF algorithm for image reconstruction from partially blocked cone-beam data. Quantitative evaluations of the reconstructed image quality were performed in the experimental studies. Results: The simulation study revealed that the developed reconstruction algorithm successfully reconstructs the images from the partial cone-beam data. In the experimental study, the proposed method effectively corrected for the scatter in each projection and reconstructed scatter-corrected images from a single scan. Reduction of cupping artifacts and an enhancement of the image contrast have been demonstrated. The image contrast has increased by a factor of about 2, and the image accuracy in terms of root-mean-square-error with respect to the fan-beam CT image has increased by more than 30%. Conclusions: The authors have successfully demonstrated that the

  12. Analytic image reconstruction from partial data for a single-scan cone-beam CT with scatter correction.

    Science.gov (United States)

    Min, Jonghwan; Pua, Rizza; Kim, Insoo; Han, Bumsoo; Cho, Seungryong

    2015-11-01

    A beam-blocker composed of multiple strips is a useful gadget for scatter correction and/or for dose reduction in cone-beam CT (CBCT). However, the use of such a beam-blocker would yield cone-beam data that can be challenging for accurate image reconstruction from a single scan in the filtered-backprojection framework. The focus of the work was to develop an analytic image reconstruction method for CBCT that can be directly applied to partially blocked cone-beam data in conjunction with the scatter correction. The authors developed a rebinned backprojection-filteration (BPF) algorithm for reconstructing images from the partially blocked cone-beam data in a circular scan. The authors also proposed a beam-blocking geometry considering data redundancy such that an efficient scatter estimate can be acquired and sufficient data for BPF image reconstruction can be secured at the same time from a single scan without using any blocker motion. Additionally, scatter correction method and noise reduction scheme have been developed. The authors have performed both simulation and experimental studies to validate the rebinned BPF algorithm for image reconstruction from partially blocked cone-beam data. Quantitative evaluations of the reconstructed image quality were performed in the experimental studies. The simulation study revealed that the developed reconstruction algorithm successfully reconstructs the images from the partial cone-beam data. In the experimental study, the proposed method effectively corrected for the scatter in each projection and reconstructed scatter-corrected images from a single scan. Reduction of cupping artifacts and an enhancement of the image contrast have been demonstrated. The image contrast has increased by a factor of about 2, and the image accuracy in terms of root-mean-square-error with respect to the fan-beam CT image has increased by more than 30%. The authors have successfully demonstrated that the proposed scanning method and image

  13. Comprehensive evaluation of ten deformable image registration algorithms for contour propagation between CT and cone-beam CT images in adaptive head & neck radiotherapy.

    Directory of Open Access Journals (Sweden)

    Xin Li

    Full Text Available Deformable image registration (DIR is a critical technic in adaptive radiotherapy (ART for propagating contours between planning computerized tomography (CT images and treatment CT/cone-beam CT (CBCT images to account for organ deformation for treatment re-planning. To validate the ability and accuracy of DIR algorithms in organ at risk (OAR contour mapping, ten intensity-based DIR strategies, which were classified into four categories-optical flow-based, demons-based, level-set-based and spline-based-were tested on planning CT and fractional CBCT images acquired from twenty-one head & neck (H&N cancer patients who underwent 6~7-week intensity-modulated radiation therapy (IMRT. Three similarity metrics, i.e., the Dice similarity coefficient (DSC, the percentage error (PE and the Hausdorff distance (HD, were employed to measure the agreement between the propagated contours and the physician-delineated ground truths of four OARs, including the vertebra (VTB, the vertebral foramen (VF, the parotid gland (PG and the submandibular gland (SMG. It was found that the evaluated DIRs in this work did not necessarily outperform rigid registration. DIR performed better for bony structures than soft-tissue organs, and the DIR performance tended to vary for different ROIs with different degrees of deformation as the treatment proceeded. Generally, the optical flow-based DIR performed best, while the demons-based DIR usually ranked last except for a modified demons-based DISC used for CT-CBCT DIR. These experimental results suggest that the choice of a specific DIR algorithm depends on the image modality, anatomic site, magnitude of deformation and application. Therefore, careful examinations and modifications are required before accepting the auto-propagated contours, especially for automatic re-planning ART systems.

  14. Comprehensive evaluation of ten deformable image registration algorithms for contour propagation between CT and cone-beam CT images in adaptive head & neck radiotherapy.

    Science.gov (United States)

    Li, Xin; Zhang, Yuyu; Shi, Yinghua; Wu, Shuyu; Xiao, Yang; Gu, Xuejun; Zhen, Xin; Zhou, Linghong

    2017-01-01

    Deformable image registration (DIR) is a critical technic in adaptive radiotherapy (ART) for propagating contours between planning computerized tomography (CT) images and treatment CT/cone-beam CT (CBCT) images to account for organ deformation for treatment re-planning. To validate the ability and accuracy of DIR algorithms in organ at risk (OAR) contour mapping, ten intensity-based DIR strategies, which were classified into four categories-optical flow-based, demons-based, level-set-based and spline-based-were tested on planning CT and fractional CBCT images acquired from twenty-one head & neck (H&N) cancer patients who underwent 6~7-week intensity-modulated radiation therapy (IMRT). Three similarity metrics, i.e., the Dice similarity coefficient (DSC), the percentage error (PE) and the Hausdorff distance (HD), were employed to measure the agreement between the propagated contours and the physician-delineated ground truths of four OARs, including the vertebra (VTB), the vertebral foramen (VF), the parotid gland (PG) and the submandibular gland (SMG). It was found that the evaluated DIRs in this work did not necessarily outperform rigid registration. DIR performed better for bony structures than soft-tissue organs, and the DIR performance tended to vary for different ROIs with different degrees of deformation as the treatment proceeded. Generally, the optical flow-based DIR performed best, while the demons-based DIR usually ranked last except for a modified demons-based DISC used for CT-CBCT DIR. These experimental results suggest that the choice of a specific DIR algorithm depends on the image modality, anatomic site, magnitude of deformation and application. Therefore, careful examinations and modifications are required before accepting the auto-propagated contours, especially for automatic re-planning ART systems.

  15. SU-F-J-81: Evaluation of Automated Deformable Registration Between Planning Computed Tomography (CT) and Daily Cone Beam CT Images Over the Course of Prostate Cancer Radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Matney, J; Hammers, J; Kaidar-Person, O; Wang, A; Chen, R; Das, S; Marks, L; Mavroidis, P [University North Carolina, Chapel Hill, NC (United States)

    2016-06-15

    Purpose: To compute daily dose delivered during radiotherapy, deformable registration needs to be relatively fast, automated, and accurate. The aim of this study was to evaluate the performance of commercial deformable registration software for deforming between two modalities: planning computed tomography (pCT) images acquired for treatment planning and cone beam (CB) CT images acquired prior to each fraction of prostate cancer radiotherapy. Methods: A workflow was designed using MIM Software™ that aligned and deformed pCT into daily CBCT images in two steps: (1) rigid shifts applied after daily CBCT imaging to align patient anatomy to the pCT and (2) normalized intensity-based deformable registration to account for interfractional anatomical variations. The physician-approved CTV and organ and risk (OAR) contours were deformed from the pCT to daily CBCT over the course of treatment. The same structures were delineated on each daily CBCT by a radiation oncologist. Dice similarity coefficient (DSC) mean and standard deviations were calculated to quantify the deformable registration quality for prostate, bladder, rectum and femoral heads. Results: To date, contour comparisons have been analyzed for 31 daily fractions of 2 of 10 of the cohort. Interim analysis shows that right and left femoral head contours demonstrate the highest agreement (DSC: 0.96±0.02) with physician contours. Additionally, deformed bladder (DSC: 0.81±0.09) and prostate (DSC: 0.80±0.07) have good agreement with physician-defined daily contours. Rectum contours have the highest variations (DSC: 0.66±0.10) between the deformed and physician-defined contours on daily CBCT imaging. Conclusion: For structures with relatively high contrast boundaries on CBCT, the MIM automated deformable registration provided accurate representations of the daily contours during treatment delivery. These findings will permit subsequent investigations to automate daily dose computation from CBCT. However

  16. SU-E-I-22: Metal Artifact Correction Using KV and Selective MV Imaging.

    Science.gov (United States)

    Keil, A; Zhu, L; Star-Lack, J; Fahrig, R

    2012-06-01

    To improve the image quality of radiotherapy planning CTs for patients with metal implants or fillings by completing the missing kV projection data with selectively acquired MV data that does not suffer from photon starvation. Using both imaging systems that are available on current radiotherapy devices, streaking artifacts are avoided and the soft tissue contrast is restored, even in areas where the kV photons do not contribute any information. This enables a better delineation of structures of interest in planning CT images for patients with metal objects. An algorithm for combining kV and MV projection data from the two on-board imagers of a radiotherapy device is presented in this work. It only requires selective MV imaging with the high energy X-rays being collimated onto the metal implants, ensuring that the patient dose does not increase significantly. The algorithm can cope with non-identical geometries of the two imagers and is based on stitching together kV and MV sinograms by estimating a ratio between them. A numerical head phantom with two dental fillings and two soft tissue patterns was used to quantitatively evaluate the proposed hybrid reconstruction algorithm. A structural similarity index (SSIM) with respect to the ground truth data was computed for two ROIs. Realistic, polychromatic spectra were used for both imagers with 120 keV(p) and 6 MeV(p). The patient dose was limited to about 6 cGy for both acquisitions combined. The reconstruction results yield visually as well as objectively better results (SSIM=74.8%) than a simple sinogram interpolation of the kV data (SSIM=69.7%) or a reconstruction from the original data (SSIM=17.9%). We have successfully implemented a new reconstruction method for hybrid kV-MV cone beam CT reconstruction that enables a better planning of radiotherapy treatments for patients with metal implants without compromising their safety. This work was funded by NIH grant 1R01CA138426-01A1. © 2012 American Association of

  17. Library based x-ray scatter correction for dedicated cone beam breast CT

    International Nuclear Information System (INIS)

    Shi, Linxi; Zhu, Lei; Vedantham, Srinivasan; Karellas, Andrew

    2016-01-01

    Purpose: The image quality of dedicated cone beam breast CT (CBBCT) is limited by substantial scatter contamination, resulting in cupping artifacts and contrast-loss in reconstructed images. Such effects obscure the visibility of soft-tissue lesions and calcifications, which hinders breast cancer detection and diagnosis. In this work, we propose a library-based software approach to suppress scatter on CBBCT images with high efficiency, accuracy, and reliability. Methods: The authors precompute a scatter library on simplified breast models with different sizes using the GEANT4-based Monte Carlo (MC) toolkit. The breast is approximated as a semiellipsoid with homogeneous glandular/adipose tissue mixture. For scatter correction on real clinical data, the authors estimate the breast size from a first-pass breast CT reconstruction and then select the corresponding scatter distribution from the library. The selected scatter distribution from simplified breast models is spatially translated to match the projection data from the clinical scan and is subtracted from the measured projection for effective scatter correction. The method performance was evaluated using 15 sets of patient data, with a wide range of breast sizes representing about 95% of general population. Spatial nonuniformity (SNU) and contrast to signal deviation ratio (CDR) were used as metrics for evaluation. Results: Since the time-consuming MC simulation for library generation is precomputed, the authors’ method efficiently corrects for scatter with minimal processing time. Furthermore, the authors find that a scatter library on a simple breast model with only one input parameter, i.e., the breast diameter, sufficiently guarantees improvements in SNU and CDR. For the 15 clinical datasets, the authors’ method reduces the average SNU from 7.14% to 2.47% in coronal views and from 10.14% to 3.02% in sagittal views. On average, the CDR is improved by a factor of 1.49 in coronal views and 2.12 in sagittal

  18. Effect of beam hardening on transmural myocardial perfusion quantification in myocardial CT imaging

    Science.gov (United States)

    Fahmi, Rachid; Eck, Brendan L.; Levi, Jacob; Fares, Anas; Wu, Hao; Vembar, Mani; Dhanantwari, Amar; Bezerra, Hiram G.; Wilson, David L.

    2016-03-01

    The detection of subendocardial ischemia exhibiting an abnormal transmural perfusion gradient (TPG) may help identify ischemic conditions due to micro-vascular dysfunction. We evaluated the effect of beam hardening (BH) artifacts on TPG quantification using myocardial CT perfusion (CTP). We used a prototype spectral detector CT scanner (Philips Healthcare) to acquire dynamic myocardial CTP scans in a porcine ischemia model with partial occlusion of the left anterior descending (LAD) coronary artery guided by pressure wire-derived fractional flow reserve (FFR) measurements. Conventional 120 kVp and 70 keV projection-based mono-energetic images were reconstructed from the same projection data and used to compute myocardial blood flow (MBF) using the Johnson-Wilson model. Under moderate LAD occlusion (FFR~0.7), we used three 5 mm short axis slices and divided the myocardium into three LAD segments and three remote segments. For each slice and each segment, we characterized TPG as the mean "endo-to-epi" transmural flow ratio (TFR). BH-induced hypoenhancement on the ischemic anterior wall at 120 kVp resulted in significantly lower mean TFR value as compared to the 70 keV TFR value (0.29+/-0.01 vs. 0.55+/-0.01 p<1e-05). No significant difference was measured between 120 kVp and 70 keV mean TFR values on segments moderately affected or unaffected by BH. In the entire ischemic LAD territory, 120 kVp mean endocardial flow was significantly reduced as compared to mean epicardial flow (15.80+/-10.98 vs. 40.85+/-23.44 ml/min/100g; p<1e-04). At 70 keV, BH was effectively minimized resulting in mean endocardial MBF of 40.85+/-15.3407 ml/min/100g vs. 74.09+/-5.07 ml/min/100g (p=0.0054) in the epicardium. We also found that BH artifact in the conventional 120 kVp images resulted in falsely reduced MBF measurements even under non-ischemic conditions.

  19. Library based x-ray scatter correction for dedicated cone beam breast CT.

    Science.gov (United States)

    Shi, Linxi; Vedantham, Srinivasan; Karellas, Andrew; Zhu, Lei

    2016-08-01

    The image quality of dedicated cone beam breast CT (CBBCT) is limited by substantial scatter contamination, resulting in cupping artifacts and contrast-loss in reconstructed images. Such effects obscure the visibility of soft-tissue lesions and calcifications, which hinders breast cancer detection and diagnosis. In this work, we propose a library-based software approach to suppress scatter on CBBCT images with high efficiency, accuracy, and reliability. The authors precompute a scatter library on simplified breast models with different sizes using the geant4-based Monte Carlo (MC) toolkit. The breast is approximated as a semiellipsoid with homogeneous glandular/adipose tissue mixture. For scatter correction on real clinical data, the authors estimate the breast size from a first-pass breast CT reconstruction and then select the corresponding scatter distribution from the library. The selected scatter distribution from simplified breast models is spatially translated to match the projection data from the clinical scan and is subtracted from the measured projection for effective scatter correction. The method performance was evaluated using 15 sets of patient data, with a wide range of breast sizes representing about 95% of general population. Spatial nonuniformity (SNU) and contrast to signal deviation ratio (CDR) were used as metrics for evaluation. Since the time-consuming MC simulation for library generation is precomputed, the authors' method efficiently corrects for scatter with minimal processing time. Furthermore, the authors find that a scatter library on a simple breast model with only one input parameter, i.e., the breast diameter, sufficiently guarantees improvements in SNU and CDR. For the 15 clinical datasets, the authors' method reduces the average SNU from 7.14% to 2.47% in coronal views and from 10.14% to 3.02% in sagittal views. On average, the CDR is improved by a factor of 1.49 in coronal views and 2.12 in sagittal views. The library-based scatter

  20. Linac-integrated 4D cone beam CT: first experimental results

    Science.gov (United States)

    Dietrich, Lars; Jetter, Siri; Tücking, Thomas; Nill, Simeon; Oelfke, Uwe

    2006-06-01

    A new online imaging approach, linac-integrated cone beam CT (CBCT), has been developed over the past few years. It has the advantage that a patient can be examined in their treatment position directly before or during a radiotherapy treatment. Unfortunately, respiratory organ motion, one of the largest intrafractional organ motions, often leads to artefacts in the reconstructed 3D images. One way to take this into account is to register the breathing phase during image acquisition for a phase-correlated image reconstruction. Therefore, the main focus of this work is to present a system which has the potential to investigate the correlation between internal (movement of the diaphragm) and external (data of a respiratory gating system) information about breathing phase and amplitude using an inline CBCT scanner. This also includes a feasibility study about using the acquired information for a respiratory-correlated 4D CBCT reconstruction. First, a moving lung phantom was used to develop and to specify the required methods which are based on an image reconstruction using only projections belonging to a certain moving phase. For that purpose, the corresponding phase has to be detected for each projection. In the case of the phantom, an electrical signal allows one to track the movement in real time. The number of projections available for the image reconstruction depends on the breathing phase and the size of the position range from which projections should be used for the reconstruction. The narrower this range is, the better the inner structures can be located, but also the noise of the images increases due to the limited number of projections. This correlation has also been analysed. In a second step, the methods were clinically applied using data sets of patients with lung tumours. In this case, the breathing phase was detected by an external gating system (AZ-733V, Anzai Medical Co.) based on a pressure sensor attached to the patient's abdominal region with a

  1. Three-dimensional anisotropic adaptive filtering of projection data for noise reduction in cone beam CT

    International Nuclear Information System (INIS)

    Maier, Andreas; Wigstroem, Lars; Hofmann, Hannes G.; Hornegger, Joachim; Zhu Lei; Strobel, Norbert; Fahrig, Rebecca

    2011-01-01

    speed-up of the processing (from 1336 to 150 s). Conclusions: Adaptive anisotropic filtering has the potential to substantially improve image quality and/or reduce the radiation dose required for obtaining 3D image data using cone beam CT.

  2. Accuracy of radiotherapy dose calculations based on cone-beam CT: comparison of deformable registration and image correction based methods

    Science.gov (United States)

    Marchant, T. E.; Joshi, K. D.; Moore, C. J.

    2018-03-01

    Radiotherapy dose calculations based on cone-beam CT (CBCT) images can be inaccurate due to unreliable Hounsfield units (HU) in the CBCT. Deformable image registration of planning CT images to CBCT, and direct correction of CBCT image values are two methods proposed to allow heterogeneity corrected dose calculations based on CBCT. In this paper we compare the accuracy and robustness of these two approaches. CBCT images for 44 patients were used including pelvis, lung and head & neck sites. CBCT HU were corrected using a ‘shading correction’ algorithm and via deformable registration of planning CT to CBCT using either Elastix or Niftyreg. Radiotherapy dose distributions were re-calculated with heterogeneity correction based on the corrected CBCT and several relevant dose metrics for target and OAR volumes were calculated. Accuracy of CBCT based dose metrics was determined using an ‘override ratio’ method where the ratio of the dose metric to that calculated on a bulk-density assigned version of the same image is assumed to be constant for each patient, allowing comparison to the patient’s planning CT as a gold standard. Similar performance is achieved by shading corrected CBCT and both deformable registration algorithms, with mean and standard deviation of dose metric error less than 1% for all sites studied. For lung images, use of deformed CT leads to slightly larger standard deviation of dose metric error than shading corrected CBCT with more dose metric errors greater than 2% observed (7% versus 1%).

  3. SU-F-J-212: Enabling Conventional Cone Beam CT with the Capability of Dual Energy Imaging Using a Simple Add-On Beam Modifier

    Energy Technology Data Exchange (ETDEWEB)

    Vinke, R; Peng, H; Xing, L [Stanford University, Palo Alto, CA (United States); Takao, S; Shirato, H [Department of Radiation Oncology, Graduate School of Medicine, Sapporo, Hokkaido (Japan); Umegaki, K [Faculty of Engineering, Hokkaido University, Sapporo, Hokkaido (Japan)

    2016-06-15

    Purpose: In searching for a robust, efficient and cost-effective dual energy cone beam CT (DECBCT) solution for various radiation oncology applications, in particularly for improved proton dose planning/replanning accuracy and DE-CBCT guided radiation therapy, we investigate a novel energy modulation scheme using a beam modifier placed between the source and patient and optimize its geometric configuration for routine clinical use. Methods: The study was performed using a Hitachi CBCT scanner and the tube voltage was set at 125 kVp. The higher energy beam was obtained by filtering the incident utilizing a beam modulation layer (material: copper, thickness: 1.8 mm). To avoid the need for double scans (one with and one without the energy modulator), the modulation layer was configured to cover only the half of the X-ray beam so that two sets of sinograms corresponding low and high energies were collected after a single gantry rotation of 360 deg. The average high energy and low energy HU numbers (HUhigh and HUlow) were derived for pixels in a defined region-of-interest, respectively. Results: The beam modifier increased the threshold of the energy spectrum from ∼20 keV up to ∼50 keV. Two complete sets of images were obtained with good alignment between the high energy and low-energy cases without any artifact observed (Fig. 2). The HUlow/HUhigh is ∼0/0 (water), ∼394/238 (brain), ∼1283/1085 (cortical bone) and ∼3000/1800 (titanium). Conclusion: The feasibility of the proposed DECT implementation using a beam modifier has been demonstrated. Compared to the existing DECT solutions, the proposed scheme is much more cost-effective and requires minimum hardware modification. The work lays foundation for us to study the quantification of HU values to derive material density images and atomic number (and electron density) of substances.

  4. TU-EF-207-05: Dedicated Cone-beam Breast CT

    International Nuclear Information System (INIS)

    Vedantham, S.

    2015-01-01

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation

  5. TU-EF-207-05: Dedicated Cone-beam Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Vedantham, S. [Univ. of Massachusetts Medical School (United States)

    2015-06-15

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation.

  6. Evaluation of Deformable Image Registration-Based Contour Propagation From Planning CT to Cone-Beam CT.

    Science.gov (United States)

    Woerner, Andrew J; Choi, Mehee; Harkenrider, Matthew M; Roeske, John C; Surucu, Murat

    2017-01-01

    We evaluated the performance of organ contour propagation from a planning computed tomography to cone-beam computed tomography with deformable image registration by comparing contours to manual contouring. Sixteen patients were retrospectively identified based on showing considerable physical change throughout the course of treatment. Multiple organs in the 3 regions (head and neck, prostate, and pancreas) were evaluated. A cone-beam computed tomography from the end of treatment was registered to the planning computed tomography using rigid registration, followed by deformable image registration. The contours were copied on cone-beam computed tomography image sets using rigid registration and modified by 2 radiation oncologists. Contours were compared using Dice similarity coefficient, mean surface distance, and Hausdorff distance. The mean physician-to-physician Dice similarity coefficient for all organs was 0.90. When compared to each physician's contours, the overall mean for rigid was 0.76 ( P cone-beam computed tomography to evaluate the changes during treatment should be used with caution.

  7. Intraoperative cone-beam CT for image-guided tibial plateau fracture reduction.

    Science.gov (United States)

    Khoury, A; Siewerdsen, J H; Whyne, C M; Daly, M J; Kreder, H J; Moseley, D J; Jaffray, D A

    2007-07-01

    A mobile isocentric C-arm was modified in our laboratory in collaboration with Siemens Medical Solutions to include a large-area flat-panel detector providing multi-mode fluoroscopy and cone-beam CT (CBCT) imaging. This technology is an important advance over existing intraoperative imaging (e.g., Iso-C(3D)), offering superior image quality, increased field of view, higher spatial resolution, and soft-tissue visibility. The aim of this study was to assess the system's performance and image quality in tibial plateau (TP) fracture reconstruction. Three TP fractures were simulated in fresh-frozen cadaveric knees through combined axial loading and lateral impact. The fractures were reduced through a lateral approach and assessed by fluoroscopy. The reconstruction was then assessed using CBCT. If necessary, further reduction and localization of remaining displaced bone fragments was performed using CBCT images for guidance. CBCT image quality was assessed with respect to projection speed, dose and filtering technique. CBCT imaging provided exquisite visualization of articular details, subtle fragment detection and localization, and confirmation of reduction and implant placement. After fluoroscopic images indicated successful initial reduction, CBCT imaging revealed areas of malalignment and displaced fragments. CBCT facilitated fragment localization and improved anatomic reduction. CBCT image noise increased gradually with reduced dose, but little difference in images resulted from increased projections. High-resolution reconstruction provided better delineation of plateau depressions. This study demonstrated a clear advantage of intraoperative CBCT over 2D fluoroscopy and Iso-C(3D) in TP fracture fixation. CBCT imaging provided benefits in fracture type diagnosis, localization of fracture fragments, and intraoperative 3D confirmation of anatomic reduction.

  8. Evaluation of robustness of maximum likelihood cone-beam CT reconstruction with total variation regularization

    International Nuclear Information System (INIS)

    Stsepankou, D; Arns, A; Hesser, J; Ng, S K; Zygmanski, P

    2012-01-01

    The objective of this paper is to evaluate an iterative maximum likelihood (ML) cone–beam computed tomography (CBCT) reconstruction with total variation (TV) regularization with respect to the robustness of the algorithm due to data inconsistencies. Three different and (for clinical application) typical classes of errors are considered for simulated phantom and measured projection data: quantum noise, defect detector pixels and projection matrix errors. To quantify those errors we apply error measures like mean square error, signal-to-noise ratio, contrast-to-noise ratio and streak indicator. These measures are derived from linear signal theory and generalized and applied for nonlinear signal reconstruction. For quality check, we focus on resolution and CT-number linearity based on a Catphan phantom. All comparisons are made versus the clinical standard, the filtered backprojection algorithm (FBP). In our results, we confirm and substantially extend previous results on iterative reconstruction such as massive undersampling of the number of projections. Errors of projection matrix parameters of up to 1° projection angle deviations are still in the tolerance level. Single defect pixels exhibit ring artifacts for each method. However using defect pixel compensation, allows up to 40% of defect pixels for passing the standard clinical quality check. Further, the iterative algorithm is extraordinarily robust in the low photon regime (down to 0.05 mAs) when compared to FPB, allowing for extremely low-dose image acquisitions, a substantial issue when considering daily CBCT imaging for position correction in radiotherapy. We conclude that the ML method studied herein is robust under clinical quality assurance conditions. Consequently, low-dose regime imaging, especially for daily patient localization in radiation therapy is possible without change of the current hardware of the imaging system. (paper)

  9. A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT)

    International Nuclear Information System (INIS)

    Liang Xin; Jacobs, Reinhilde; Hassan, Bassam; Li Limin; Pauwels, Ruben; Corpas, Livia; Souza, Paulo Couto; Martens, Wendy; Shahbazian, Maryam; Alonso, Arie

    2010-01-01

    Aims: To compare image quality and visibility of anatomical structures in the mandible between five Cone Beam Computed Tomography (CBCT) scanners and one Multi-Slice CT (MSCT) system. Materials and methods: One dry mandible was scanned with five CBCT scanners (Accuitomo 3D, i-CAT, NewTom 3G, Galileos, Scanora 3D) and one MSCT system (Somatom Sensation 16) using 13 different scan protocols. Visibility of 11 anatomical structures and overall image noise were compared between CBCT and MSCT. Five independent observers reviewed the CBCT and the MSCT images in the three orthographic planes (axial, sagittal and coronal) and assessed image quality on a five-point scale. Results: Significant differences were found in the visibility of the different anatomical structures and image noise level between MSCT and CBCT and among the five CBCT systems (p = 0.0001). Delicate structures such as trabecular bone and periodontal ligament were significantly less visible and more variable among the systems in comparison with other anatomical structures (p = 0.0001). Visibility of relatively large structures such as mandibular canal and mental foramen was satisfactory for all devices. The Accuitomo system was superior to MSCT and all other CBCT systems in depicting anatomical structures while MSCT was superior to all other CBCT systems in terms of reduced image noise. Conclusions: CBCT image quality is comparable or even superior to MSCT even though some variability exists among the different CBCT systems in depicting delicate structures. Considering the low radiation dose and high-resolution imaging, CBCT could be beneficial for dentomaxillofacial radiology.

  10. A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT)

    Energy Technology Data Exchange (ETDEWEB)

    Liang Xin, E-mail: Xin.Liang@med.kuleuven.b [Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); College of Stomatology, Dalian Medical University (China); Jacobs, Reinhilde, E-mail: Reinhilde.Jacobs@uz.kuleuven.b [Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); Hassan, Bassam, E-mail: b.hassan@acta.n [Department of Oral Radiology, Academic Centre for Dentistry Amsterdam (ACTA), Amsterdam (Netherlands); Li Limin, E-mail: Limin.Li@uz.kuleuven.b [Department of Paediatric Dentistry and Special Dental Care, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); Pauwels, Ruben, E-mail: Ruben.Pauwels@med.kuleuven.b [Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); Corpas, Livia, E-mail: LiviaCorpas@gmail.co [Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); Souza, Paulo Couto, E-mail: Paulo.CoutoSouza@med.kuleuven.b [Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); Martens, Wendy, E-mail: wendy.martens@uhasselt.b [Department of Basic Medical Sciences, Faculty of Medicine, University of Hasselt, Diepenbeek (Belgium); Shahbazian, Maryam, E-mail: Maryam.Shahbazian@student.kuleuven.b [Oral Imaging Centre, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Faculty of Medicine, Catholic University of Leuven (Belgium); Alonso, Arie, E-mail: ariel.alonso@uhasselt.b [Department of Biostatistics and Statistical Bioinformatics, Universiteit Hasselt (Belgium)

    2010-08-15

    Aims: To compare image quality and visibility of anatomical structures in the mandible between five Cone Beam Computed Tomography (CBCT) scanners and one Multi-Slice CT (MSCT) system. Materials and methods: One dry mandible was scanned with five CBCT scanners (Accuitomo 3D, i-CAT, NewTom 3G, Galileos, Scanora 3D) and one MSCT system (Somatom Sensation 16) using 13 different scan protocols. Visibility of 11 anatomical structures and overall image noise were compared between CBCT and MSCT. Five independent observers reviewed the CBCT and the MSCT images in the three orthographic planes (axial, sagittal and coronal) and assessed image quality on a five-point scale. Results: Significant differences were found in the visibility of the different anatomical structures and image noise level between MSCT and CBCT and among the five CBCT systems (p = 0.0001). Delicate structures such as trabecular bone and periodontal ligament were significantly less visible and more variable among the systems in comparison with other anatomical structures (p = 0.0001). Visibility of relatively large structures such as mandibular canal and mental foramen was satisfactory for all devices. The Accuitomo system was superior to MSCT and all other CBCT systems in depicting anatomical structures while MSCT was superior to all other CBCT systems in terms of reduced image noise. Conclusions: CBCT image quality is comparable or even superior to MSCT even though some variability exists among the different CBCT systems in depicting delicate structures. Considering the low radiation dose and high-resolution imaging, CBCT could be beneficial for dentomaxillofacial radiology.

  11. Image-based motion compensation for high-resolution extremities cone-beam CT

    Science.gov (United States)

    Sisniega, A.; Stayman, J. W.; Cao, Q.; Yorkston, J.; Siewerdsen, J. H.; Zbijewski, W.

    2016-03-01

    Purpose: Cone-beam CT (CBCT) of the extremities provides high spatial resolution, but its quantitative accuracy may be challenged by involuntary sub-mm patient motion that cannot be eliminated with simple means of external immobilization. We investigate a two-step iterative motion compensation based on a multi-component metric of image sharpness. Methods: Motion is considered with respect to locally rigid motion within a particular region of interest, and the method supports application to multiple locally rigid regions. Motion is estimated by maximizing a cost function with three components: a gradient metric encouraging image sharpness, an entropy term that favors high contrast and penalizes streaks, and a penalty term encouraging smooth motion. Motion compensation involved initial coarse estimation of gross motion followed by estimation of fine-scale displacements using high resolution reconstructions. The method was evaluated in simulations with synthetic motion (1-4 mm) applied to a wrist volume obtained on a CMOS-based CBCT testbench. Structural similarity index (SSIM) quantified the agreement between motion-compensated and static data. The algorithm was also tested on a motion contaminated patient scan from dedicated extremities CBCT. Results: Excellent correction was achieved for the investigated range of displacements, indicated by good visual agreement with the static data. 10-15% improvement in SSIM was attained for 2-4 mm motions. The compensation was robust against increasing motion (4% decrease in SSIM across the investigated range, compared to 14% with no compensation). Consistent performance was achieved across a range of noise levels. Significant mitigation of artifacts was shown in patient data. Conclusion: The results indicate feasibility of image-based motion correction in extremities CBCT without the need for a priori motion models, external trackers, or fiducials.

  12. Correction of patient motion in cone-beam CT using 3D-2D registration

    Science.gov (United States)

    Ouadah, S.; Jacobson, M.; Stayman, J. W.; Ehtiati, T.; Weiss, C.; Siewerdsen, J. H.

    2017-12-01

    Cone-beam CT (CBCT) is increasingly common in guidance of interventional procedures, but can be subject to artifacts arising from patient motion during fairly long (~5-60 s) scan times. We present a fiducial-free method to mitigate motion artifacts using 3D-2D image registration that simultaneously corrects residual errors in the intrinsic and extrinsic parameters of geometric calibration. The 3D-2D registration process registers each projection to a prior 3D image by maximizing gradient orientation using the covariance matrix adaptation-evolution strategy optimizer. The resulting rigid transforms are applied to the system projection matrices, and a 3D image is reconstructed via model-based iterative reconstruction. Phantom experiments were conducted using a Zeego robotic C-arm to image a head phantom undergoing 5-15 cm translations and 5-15° rotations. To further test the algorithm, clinical images were acquired with a CBCT head scanner in which long scan times were susceptible to significant patient motion. CBCT images were reconstructed using a penalized likelihood objective function. For phantom studies the structural similarity (SSIM) between motion-free and motion-corrected images was  >0.995, with significant improvement (p  values of uncorrected images. Additionally, motion-corrected images exhibited a point-spread function with full-width at half maximum comparable to that of the motion-free reference image. Qualitative comparison of the motion-corrupted and motion-corrected clinical images demonstrated a significant improvement in image quality after motion correction. This indicates that the 3D-2D registration method could provide a useful approach to motion artifact correction under assumptions of local rigidity, as in the head, pelvis, and extremities. The method is highly parallelizable, and the automatic correction of residual geometric calibration errors provides added benefit that could be valuable in routine use.

  13. Comparison of panoramic radiography and cone beam CT in the assessment of juxta-apical radiolucency.

    Science.gov (United States)

    Nascimento, Eduarda Helena Leandro; Oenning, Anne Caroline Costa; Freire, Bernardo Barbosa; Gaêta-Araujo, Hugo; Haiter-Neto, Francisco; Freitas, Deborah Queiroz

    2018-01-01

    To compare the performance of panoramic radiography (PAN) and cone beam CT (CBCT) in the detection of juxta-apical radiolucency (JAR), as well as to investigate, in CBCT images, if there are factors associated with the detection of JAR on PAN. Two oral radiologists assessed the presence of JAR in PAN and CBCT images of 175 individuals (308 mandibular third molars). The cortical plates involvement and the JAR size and location were assessed on CBCT to evaluate if these factors were related to JAR detection on PAN. McNemar's test and multiple logistic regression were performed. PAN and CBCT differed significantly in the detection of JAR (p = 0.001). On PAN, JAR was identified on 24% of the patients while on CBCT its detection increased to 32.6%. JAR was detected only on CBCT and only on PAN in 26 and 7 cases, respectively. Distal/mesial surfaces of dental roots were where JAR was mostly located (84.5%), cortical thinning was found in 59.2% of cases and the mean (SD) of JAR size was 5.03 (±1.8) mm. However, these factors were not associated with JAR detection on PAN (p > 0.05). On the other hand, the location of the cortical involvement (if buccal or lingual) was associated with JAR detection on PAN, which was more detectable when the thinning was on buccal cortical. Juxta-apical radiolucency is more often detected on CBCT than on PAN. JAR detection on PAN was improved when it was related to the buccal cortical plate of the mandible.

  14. A method for robust segmentation of arbitrarily shaped radiopaque structures in cone-beam CT projections.

    Science.gov (United States)

    Poulsen, Per Rugaard; Fledelius, Walther; Keall, Paul J; Weiss, Elisabeth; Lu, Jun; Brackbill, Emily; Hugo, Geoffrey D

    2011-04-01

    Implanted markers are commonly used in radiotherapy for x-ray based target localization. The projected marker position in a series of cone-beam CT (CBCT) projections can be used to estimate the three dimensional (3D) target trajectory during the CBCT acquisition. This has important applications in tumor motion management such as motion inclusive, gating, and tumor tracking strategies. However, for irregularly shaped markers, reliable segmentation is challenged by large variations in the marker shape with projection angle. The purpose of this study was to develop a semiautomated method for robust and reliable segmentation of arbitrarily shaped radiopaque markers in CBCT projections. The segmentation method involved the following three steps: (1) Threshold based segmentation of the marker in three to six selected projections with large angular separation, good marker contrast, and uniform background; (2) construction of a 3D marker model by coalignment and backprojection of the threshold-based segmentations; and (3) construction of marker templates at all imaging angles by projection of the 3D model and use of these templates for template-based segmentation. The versatility of the segmentation method was demonstrated by segmentation of the following structures in the projections from two clinical CBCT scans: (1) Three linear fiducial markers (Visicoil) implanted in or near a lung tumor and (2) an artificial cardiac valve in a lung cancer patient. Automatic marker segmentation was obtained in more than 99.9% of the cases. The segmentation failed in a few cases where the marker was either close to a structure of similar appearance or hidden behind a dense structure (data cable). A robust template-based method for segmentation of arbitrarily shaped radiopaque markers in CBCT projections was developed.

  15. [Di Paolo's cephalometrical analysis of lower face by means of Cone-Beam CT].

    Science.gov (United States)

    Dobai, Adrienn; Vizkelety, Tamás; Markella, Zsolt; Rosta, Adrienne; Kucserá, Ágnes; Barabás, József

    2016-06-01

    3D cephalometry is often the only way to set up accurate diagnosis and treatment plan in the field of reconstructive surgery. In these cases complement exposures are needed beyond common cephalograms with higher accuracy than conventional Cone-Beam CT. Consequently the aim of our study was to perform a complex 3D cephalometry. As the first step of this approach, was the 3D adaptation of DiPaolo's Quadrilateral technique, and to determine norms of references in lower face by means of CBCT. Thirty non-orthodontic CBCT scans were selected for the digitalization. The most important inclusion criteria was Class I occlusion. Locations of 55 landmarks were signed three times by three observers by means of Cranio Viewer software. However Quadrilateral analysis contains only millimetric values we also integrate angles in the 3D version to determine the width of maxilla and mandible. In the 2D examination--where landmarks were projected to the middle plane. The SDs of the lengths were between 2,66 mm and 5,20 mm. The ratios of normodivergent lower face were significant different from the one by DiPaolo. In 3D adaptation there were no significant differences between the measurements of the two sides (p ≥ 0.05). We found mostly strong and significant correlations between each anatomical structure except of angles. Creation of 3D Quadrilateral cephalometry by means of strong correlation and norms of Class I occlusion provide a practical, reliable method to measure also the transversal asymmetry of lower face which is necessary part of 3D cephalometry.

  16. Follow-up of aortic intramural hematoma by electron beam CT

    International Nuclear Information System (INIS)

    Zhi Aihua; Dai Ruping; Jiang Shiliang; Jin Jinglin; Chao Cehng; Bao Hua; Wu Haiying; Sun Lizhong

    2006-01-01

    Objective: To evaluate the dynamic changing and prognosis of aortic intramural hematoma (IMH) by electron beam CT(EBCT). Methods: A retrospective study was performed during a period of time from January. 2002 to September 2004. Twenty-three patients with aortic IMH were scanned by EBCT more than 2 times. The patients were followed-up for 4-405 days (mean 105.8 days). Mean follow-up was 105.84 days. In this group 23 patients (19 men and 4 women, aged from 30 to 81 years, mean 57.78 years) who had been undergone EBCT were diagnosed with IMH. EBCT scanner was used with a model of Imatron C-150XP. Contrast-enhanced continuous volume scanning (CVS) was performed. The slice thickness was 6 mm or 3 mm with an acquisition time of 100 milliseconds. A total amount of contrast media (Ultravist 300 or 320, or Ominpaque 300 or 320 mg/ml) of 80-100 ml was used with the rate of 3.5- 4.5 ml/s. The scan delayed time was 18-30 s. Results: Four patients with Standford A IMH were diagnosed. Among them, 3 patients were observed in complete regression, 1 patient was observed without changing. Nineteen patients with Standford B IMH were diagnosed. Among them, 9 patients were observed in complete regression, 4 patients were observed in incomplete regression, 4 patients were observed without changing, 2 patients were observed in progression on EBCT. From 4 days to 15 days, IMH was observed without change or worse. From 16 days to 30 days IMH was observed in regression. Conclusion: EBCT was a very useful tool for detecting and following up of IMH. The different features of the involved aortic walls shown on EBCT were used for planning surgery. (authors)

  17. Electron-beam CT diagnosis of the viscero-atrial heterotaxy syndrome

    International Nuclear Information System (INIS)

    Yang Youyou; Dai Ruping; Jing Baolian; He Sha; Bai Hua; Li Xiangmin; Zhou Xuhui; Peng Qian; Meng Quanfei

    2002-01-01

    Objective: To assess the usefulness of electron-beam CT (EBCT) in diagnosis of the viscero-atrial heterotaxy syndrome. Methods: Ten patients with the viscero-atrial heterotaxy syndrome were evaluated. The patients ranged in age from 7 months to 17 years (averaged 9.5 years). Five of the patients underwent EBCT contrast single slice mode while another 5 patients did continuous volume scan obtained from the superior aperture of thorax to the middle part of abdomen. All the patients had both angiocardiogram and echocardiogram, and 6 patients had operative outcomes. Results; (1) Eight patients with right atrial isomerism, bilaterally morphologic right atrial appendages, right lobe dominant symmetric liver, bilaterally epi-arterial bronchi, trilobed lungs, and total anomalous pulmonary venous connectional were clearly detected. Endocardial cushions defect, pulmonary stenosis, right-sided aortic arch and descending aorta were documented in 7 patients. Double outlet of right ventricle was imaged in 6 patients and hiatal hernia in 2 patients. Neither a spleen nor splenulus were found. (2) Two patients with left atrial isomerism, bilaterally morphologic left atrial appendages, left lobe dominant symmetric liver, bilaterally hyparterial bronchi, bilobed lungs, double outlet of right ventricle, pulmonary stenosis, interruption of inferior vena cave, right-sided aortic arch, and descending aorta were documented. Endocardial cushions defect was detected in 1 patient. Multiple spleens were demonstrated in the right upper abdomen in the 2 patients. (3) The number of abnormal observations detected by EBCT was 116, while that done by angiocardiogram and echocardiogram were 65 and 43 respectively. Conclusion: EBCT is a useful tool in the evaluation of patients with the syndrome of viscero-atrial heterotaxia

  18. Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

    Science.gov (United States)

    Al-Okshi, A; Lindh, C; Salé, H; Gunnarsson, M; Rohlin, M

    2015-01-01

    To estimate effective dose of cone beam CT (CBCT) of the facial skeleton with focus on measurement methods and scanning protocols. A systematic review, which adhered to the preferred reporting items for systematic reviews (PRISMA) Statement, of the literature up to April 2014 was conducted. Data sources included MEDLINE®, The Cochrane Library and Web of Science. A model was developed to underpin data extraction from 38 included studies. Technical specifications of the CBCT units were insufficiently described. Heterogeneity in measurement methods and scanning protocols between studies made comparisons of effective doses of different CBCT units and scanning protocols difficult. Few studies related doses to image quality. Reported effective dose varied across studies, ranging between 9.7 and 197.0 μSv for field of views (FOVs) with height ≤5 cm, between 3.9 and 674.0 μSv for FOVs of heights 5.1-10.0 cm and between 8.8 and 1073.0 μSv for FOVs >10 cm. There was an inconsistency regarding reported effective dose of studies of the same CBCT unit with the same FOV dimensions. The review reveals a need for studies on radiation dosages related to image quality. Reporting quality of future studies has to be improved to facilitate comparison of effective doses obtained from examinations with different CBCT units and scanning protocols. A model with minimum data set on important parameters based on this observation is proposed. Data important when estimating effective dose were insufficiently reported in most studies. A model with minimum data based on this observation is proposed. Few studies related effective dose to image quality.

  19. WE-G-18A-05: Cone-Beam CT Reconstruction with Deformed Prior Image

    International Nuclear Information System (INIS)

    Zhang, H; Huang, J; Ma, J; Chen, W; Ouyang, L; Wang, J

    2014-01-01

    Purpose: Prior image can be incorporated into image reconstruction process to improve the quality of on-treatment cone-beam CT (CBCT) from sparseview or low-dose projections. However, the deformation between the prior image and on-treatment CBCT are not considered in current prior image based reconstructions (e.g., prior image constrained compressed sensing (PICCS)). The purpose of this work is to develop a deformed-prior-imagebased- reconstruction strategy (DPIR) to address the mismatch problem between the prior image and target image. Methods: The deformed prior image is obtained by a projection based registration approach. Specifically, the deformation vector fields (DVF) used to deform the prior image is estimated through matching the forward projection of the prior image and the measured on-treatment projection. The deformed prior image is then used as the prior image in the standard PICCS algorithm. Simulation studies on the XCAT phantom was conducted to evaluate the performance of the projection based registration procedure and the proposed DPIR strategy. Results: The deformed prior image matches the geometry of on-treatment CBCT closer as compared to the original prior image. Using the deformed prior image, the quality of the image reconstructed by DPIR from few-view projection data is greatly improved as compared to the standard PICCS algorithm. The relative image reconstruction error is reduced to 11.13% in the proposed DPIR from 17.57% in the original PICCS. Conclusion: The proposed DPIR approach can solve the mismatch problem between the prior image and target image, which overcomes the limitation of the original PICCS algorithm for CBCT reconstruction from sparse-view or low-dose projections

  20. Few-view cone-beam CT reconstruction with deformed prior image

    International Nuclear Information System (INIS)

    Zhang, Hua; Ouyang, Luo; Wang, Jing; Huang, Jing; Ma, Jianhua; Chen, Wufan

    2014-01-01

    Purpose: Prior images can be incorporated into the image reconstruction process to improve the quality of subsequent cone-beam CT (CBCT) images from sparse-view or low-dose projections. The purpose of this work is to develop a deformed prior image-based reconstruction (DPIR) strategy to mitigate the deformation between the prior image and the target image. Methods: The deformed prior image is obtained by a projection-based registration approach. Specifically, the deformation vector fields used to deform the prior image are estimated through iteratively matching the forward projection of the deformed prior image and the measured on-treatment projections. The deformed prior image is then used as the prior image in the standard prior image constrained compressed sensing (PICCS) algorithm. A simulation study on an XCAT phantom and a clinical study on a head-and-neck cancer patient were conducted to evaluate the performance of the proposed DPIR strategy. Results: The deformed prior image matches the geometry of the on-treatment CBCT more closely as compared to the original prior image. Consequently, the performance of the DPIR strategy from few-view projections is improved in comparison to the standard PICCS algorithm, based on both visual inspection and quantitative measures. In the XCAT phantom study using 20 projections, the average root mean squared error is reduced from 14% in PICCS to 10% in DPIR, and the average universal quality index increases from 0.88 in PICCS to 0.92 in DPIR. Conclusions: The present DPIR approach provides a practical solution to the mismatch problem between the prior image and target image, which improves the performance of the original PICCS algorithm for CBCT reconstruction from few-view or low-dose projections

  1. Calibration of megavoltage cone-beam CT for radiotherapy dose calculations: Correction of cupping artifacts and conversion of CT numbers to electron density

    International Nuclear Information System (INIS)

    Petit, Steven F.; Elmpt, Wouter J. C. van; Nijsten, Sebastiaan M. J. J. G.; Lambin, Philippe; Dekker, Andre L. A. J.

    2008-01-01

    Megavoltage cone-beam CT (MV CBCT) is used for three-dimensional imaging of the patient anatomy on the treatment table prior to or just after radiotherapy treatment. To use MV CBCT images for radiotherapy dose calculation purposes, reliable electron density (ED) distributions are needed. Patient scatter, beam hardening and softening effects result in cupping artifacts in MV CBCT images and distort the CT number to ED conversion. A method based on transmission images is presented to correct for these effects without using prior knowledge of the object's geometry. The scatter distribution originating from the patient is calculated with pencil beam scatter kernels that are fitted based on transmission measurements. The radiological thickness is extracted from the scatter subtracted transmission images and is then converted to the primary transmission used in the cone-beam reconstruction. These corrections are performed in an iterative manner, without using prior knowledge regarding the geometry and composition of the object. The method was tested using various homogeneous and inhomogeneous phantoms with varying shapes and compositions, including a phantom with different electron density inserts, phantoms with large density variations, and an anthropomorphic head phantom. For all phantoms, the cupping artifact was substantially removed from the images and a linear relation between the CT number and electron density was found. After correction the deviations in reconstructed ED from the true values were reduced from up to 0.30 ED units to 0.03 for the majority of the phantoms; the residual difference is equal to the amount of noise in the images. The ED distributions were evaluated in terms of absolute dose calculation accuracy for homogeneous cylinders of different size; errors decreased from 7% to below 1% in the center of the objects for the uncorrected and corrected images, respectively, and maximum differences were reduced from 17% to 2%, respectively. The

  2. Comparison of prostate positioning guided by three-dimensional transperineal ultrasound and cone beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Li, Minglun; Ballhausen, Hendrik; Hegemann, Nina-Sophie; Reiner, Michael; Manapov, Farkhad; Corradini, Stefanie; Ganswindt, Ute; Belka, Claus [University Hospital Munich, LMU Munich, Department of Radiation Oncology, Munich (Germany); Tritschler, Stefan; Gratzke, Christian [University Hospital Munich, LMU Munich, Department of Urology, Munich (Germany)

    2017-03-15

    The accuracy of a transperineal three-dimensional ultrasound system (3DUS) was assessed for prostate positioning and compared to fiducial- and bone-based positioning in kV cone beam computed tomography (CBCT) during definitive radiotherapy of prostate cancer. Each of the 7 patients had three fiducial markers implanted into the prostate before treatment. Prostate positioning was simultaneously measured by 3DUS and CBCT before each fraction. In total, 177 pairs of 3DUS and CBCT scans were collected. Bone-match and seed-match were performed for each CBCT. Using seed-match as a reference, the accuracy of 3DUS and bone-match was evaluated. Systematic and random errors as well as optimal setup margins were calculated for 3DUS and bone-match. The discrepancy between 3DUS and seed-match in CBCT (average ± standard deviation) was 0.0 ± 1.7 mm laterally, 0.2 ± 2.0 mm longitudinally, and 0.3 ± 1.7 mm vertically. Using seed-match as a reference, systematic errors for 3DUS were 1.2 mm, 1.1 mm, and 0.9 mm; and random errors were 1.4 mm, 1.8 mm, and 1.6 mm, on lateral, longitudinal, and vertical axes, respectively. By analogy, the difference of bone-match to seed-match was 0.1 ± 1.1 mm laterally, 1.3 ± 3.8 mm longitudinally, and 1.3 ± 4.5 mm vertically. Systematic errors were 0.5 mm, 2.2 mm, and 2.6 mm; and random errors were 1.0 mm, 3.1 mm, and 3.9 mm on lateral, longitudinal, and vertical axes, respectively. The accuracy of 3DUS was significantly higher than that of bone-match on longitudinal and vertical axes, but not on the lateral axis. Image-guided radiotherapy of prostate cancer based on transperineal 3DUS was feasible, with overall small discrepancy to seed-match in CBCT in this retrospective study. Compared to bone-match, transperineal 3DUS achieved higher accuracy on longitudinal and vertical axes. (orig.) [German] Bewertung der Genauigkeit eines transperinealen dreidimensionalen Ultraschallsystems (3DUS) fuer die Prostatapositionierung und Vergleich mit

  3. CT of small objects with synchrotron radiation from the X-26C beam line at the National Synchrotron Light Source

    International Nuclear Information System (INIS)

    Spanne, P.; Rivers, M.L.

    1987-01-01

    CT with spatial resolution in the micrometer range has so far not been possible because of the limited photon fluence rates available from conventional x-ray sources. Synchrotron-generated x-rays now make such high-resolution imaging possible. Experiments to develop CT to the micrometer spatial resolution range have been performed at the X-26C microprobe beam line at the National Synchrotron Light Source at Brookhaven National Laboratory. They have so far yielded images with spatial resolutions down to about 20 μm and show that there is a potential to improve the spatial resolution down to at least 1 μm in imaging of dead objects. This creates an entirely new tool for imaging of microstructures in needle biopsies. Calculations shown that the choice of photon energy is critical in minimizing radiation damage and the imaging time. They also show that CT imaging of small laboratory animals, for example, mice and rats, cna be done with a spatial resolution of the order of 50 μm without inducing any significant radiation damage in the animals. By use of a line-shaped x-ray beam, a photodiode array detector, and standard CAMAC acquisition modules, it should be possible to obtain an image in about 1 minute

  4. Evaluation of the effective dose of cone beam CT and multislice CT for temporomandibular joint examinations at optimized exposure levels.

    Science.gov (United States)

    Kadesjö, N; Benchimol, D; Falahat, B; Näsström, K; Shi, X-Q

    2015-01-01

    To compare the effective dose to patients from temporomandibular joint examinations using a dental CBCT device and a multislice CT (MSCT) device, both before and after dose optimization. A Promax(®) 3D (Planmeca, Helsinki, Finland) dental CBCT and a LightSpeed VCT(®) (GE Healthcare, Little Chalfont, UK) multislice CT were used. Organ doses and effective doses were estimated from thermoluminescent dosemeters at 61 positions inside an anthropomorphic phantom at the exposure settings in clinical use. Optimized exposure protocols were obtained through an optimization study using a dry skull phantom, where four observers rated image quality taken at different exposure levels. The optimal exposure level was obtained when all included criteria were rated as acceptable or better by all observers. The effective dose from a bilateral examination was 184 µSv for Promax 3D and 113 µSv for LightSpeed VCT before optimization. Post optimization, the bilateral effective dose was 92 µSv for Promax 3D and 124 µSv for LightSpeed VCT. At optimized exposure levels, the effective dose from CBCT was comparable to MSCT.

  5. SU-D-206-03: Segmentation Assisted Fast Iterative Reconstruction Method for Cone-Beam CT

    International Nuclear Information System (INIS)

    Wu, P; Mao, T; Gong, S; Wang, J; Niu, T; Sheng, K; Xie, Y

    2016-01-01

    Purpose: Total Variation (TV) based iterative reconstruction (IR) methods enable accurate CT image reconstruction from low-dose measurements with sparse projection acquisition, due to the sparsifiable feature of most CT images using gradient operator. However, conventional solutions require large amount of iterations to generate a decent reconstructed image. One major reason is that the expected piecewise constant property is not taken into consideration at the optimization starting point. In this work, we propose an iterative reconstruction method for cone-beam CT (CBCT) using image segmentation to guide the optimization path more efficiently on the regularization term at the beginning of the optimization trajectory. Methods: Our method applies general knowledge that one tissue component in the CT image contains relatively uniform distribution of CT number. This general knowledge is incorporated into the proposed reconstruction using image segmentation technique to generate the piecewise constant template on the first-pass low-quality CT image reconstructed using analytical algorithm. The template image is applied as an initial value into the optimization process. Results: The proposed method is evaluated on the Shepp-Logan phantom of low and high noise levels, and a head patient. The number of iterations is reduced by overall 40%. Moreover, our proposed method tends to generate a smoother reconstructed image with the same TV value. Conclusion: We propose a computationally efficient iterative reconstruction method for CBCT imaging. Our method achieves a better optimization trajectory and a faster convergence behavior. It does not rely on prior information and can be readily incorporated into existing iterative reconstruction framework. Our method is thus practical and attractive as a general solution to CBCT iterative reconstruction. This work is supported by the Zhejiang Provincial Natural Science Foundation of China (Grant No. LR16F010001), National High-tech R

  6. A One-Step Cone-Beam CT-Enabled Planning-to-Treatment Model for Palliative Radiotherapy-From Development to Implementation

    International Nuclear Information System (INIS)

    Wong, Rebecca K.S.; Letourneau, Daniel; Varma, Anita; Bissonnette, Jean Pierre; Fitzpatrick, David; Grabarz, Daniel; Elder, Christine; Martin, Melanie; Bezjak, Andrea; Panzarella, Tony; Gospodarowicz, Mary; Jaffray, David A.

    2012-01-01

    Purpose: To develop a cone-beam computed tomography (CT)–enabled one-step simulation-to-treatment process for the treatment of bone metastases. Methods and Materials: A three-phase prospective study was conducted. Patients requiring palliative radiotherapy to the spine, mediastinum, or abdomen/pelvis suitable for treatment with simple beam geometry (≤2 beams) were accrued. Phase A established the accuracy of cone-beam CT images for the purpose of gross tumor target volume (GTV) definition. Phase B evaluated the feasibility of implementing the cone-beam CT–enabled planning process at the treatment unit. Phase C evaluated the online cone-beam CT–enabled process for the planning and treatment of patients requiring radiotherapy for bone metastases. Results: Eighty-four patients participated in this study. Phase A (n = 9) established the adequacy of cone-beam CT images for target definition. Phase B (n = 45) established the quality of treatment plans to be adequate for clinical implementation for bone metastases. When the process was applied clinically in bone metastases (Phase C), the degree of overlap between planning computed tomography (PCT) and cone-beam CT for GTV and between PCT and cone-beam CT for treatment field was 82% ± 11% and 97% ± 4%, respectively. The oncologist’s decision to accept the plan under a time-pressured environment remained of high quality, with the cone-beam CT–generated treatment plan delivering at least 90% of the prescribed dose to 100% ± 0% of the cone-beam CT planning target volume (PTV). With the assumption that the PCT PTV is the gold-standard target, the cone-beam CT–generated treatment plan delivered at least 90% and at least 95% of dose to 98% ± 2% and 97% ± 5% of the PCT PTV, respectively. The mean time for the online planning and treatment process was 32.7 ± 4.0 minutes. Patient satisfaction was high, with a trend for superior satisfaction with the cone-beam CT–enabled process. Conclusions: The cone-beam CT

  7. Diagnostic accuracy of cone-beam CT in the assessment of mandibular invasion of lower gingival carcinoma: Comparison with conventional panoramic radiography

    Energy Technology Data Exchange (ETDEWEB)

    Momin, Mohammad A. [Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: momin.orad@tmd.ac.jp; Okochi, Kiyoshi [Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: kiyoshi.orad@tmd.ac.jp; Watanabe, Hiroshi [Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: hiro.orad@tmd.ac.jp; Imaizumi, Akiko [Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: ima.orad@tmd.ac.jp; Omura, Ken [Oral Surgery, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: omura.osur@tmd.ac.jp; Amagasa, Teruo [Maxillofacial Surgery, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: t-amagasa.mfs@tmd.ac.jp; Okada, Norihiko [Diagnostic Oral Pathology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: nokd.opth@tmd.ac.jp; Ohbayashi, Naoto [Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: nao.orad@tmd.ac.jp; Kurabayashi, Tohru [Oral and Maxillofacial Radiology, Graduate School, Tokyo Medical and Dental University, Yushima 1-5-45, Bunkyo-ku, Tokyo 113-8549 (Japan)], E-mail: kura.orad@tmd.ac.jp

    2009-10-15

    Purpose: To evaluate the diagnostic accuracy of cone-beam CT in assessing mandibular invasion by lower gingival carcinoma and compare it with that of panoramic radiography. Patients and methods: Fifty patients with squamous cell carcinoma of the lower gingiva who were examined by both panoramic radiography and cone-beam CT before surgery were included in this study. Five radiologists used a 6-point rating scale to independently evaluate cone-beam CT and panoramic images for the presence or absence of alveolar bone and mandibular canal involvement by tumor. Using the histopathogical findings as the gold standard, we calculated and compared the area under the receiver operating characteristic curve (Az value) and the sensitivity and specificity of the two imaging modalities. Results: In evaluations of both alveolar bone and mandibular canal involvement, the mean Az value for cone-beam CT (0.918 and 0.977, respectively) was significantly higher than that for panoramic radiography (0.793 and 0.872, respectively). The mean sensitivity for cone-beam CT (89% and 99%, respectively) was significantly higher than that for panoramic radiography (73% and 56%, respectively). There was no significant difference in the mean specificity. While cone-beam CT could provide high-resolution three-dimensional images, the image quality around the alveolar crest was often hampered by severe dental artifacts and image noise, resulting in difficulties in detecting subtle alveolar invasion. Conclusion: Cone-beam CT was significantly superior to panoramic radiography in evaluating mandibular invasion by lower gingival carcinoma. Its diagnostic value in detecting subtle alveolar invasion, however, may be limited by severe dental artifacts and image noise.

  8. Diagnostic accuracy of cone-beam CT in the assessment of mandibular invasion of lower gingival carcinoma: comparison with conventional panoramic radiography.

    Science.gov (United States)

    Momin, Mohammad A; Okochi, Kiyoshi; Watanabe, Hiroshi; Imaizumi, Akiko; Omura, Ken; Amagasa, Teruo; Okada, Norihiko; Ohbayashi, Naoto; Kurabayashi, Tohru

    2009-10-01

    To evaluate the diagnostic accuracy of cone-beam CT in assessing mandibular invasion by lower gingival carcinoma and compare it with that of panoramic radiography. Fifty patients with squamous cell carcinoma of the lower gingiva who were examined by both panoramic radiography and cone-beam CT before surgery were included in this study. Five radiologists used a 6-point rating scale to independently evaluate cone-beam CT and panoramic images for the presence or absence of alveolar bone and mandibular canal involvement by tumor. Using the histopathogical findings as the gold standard, we calculated and compared the area under the receiver operating characteristic curve (Az value) and the sensitivity and specificity of the two imaging modalities. In evaluations of both alveolar bone and mandibular canal involvement, the mean Az value for cone-beam CT (0.918 and 0.977, respectively) was significantly higher than that for panoramic radiography (0.793 and 0.872, respectively). The mean sensitivity for cone-beam CT (89% and 99%, respectively) was significantly higher than that for panoramic radiography (73% and 56%, respectively). There was no significant difference in the mean specificity. While cone-beam CT could provide high-resolution three-dimensional images, the image quality around the alveolar crest was often hampered by severe dental artifacts and image noise, resulting in difficulties in detecting subtle alveolar invasion. Cone-beam CT was significantly superior to panoramic radiography in evaluating mandibular invasion by lower gingival carcinoma. Its diagnostic value in detecting subtle alveolar invasion, however, may be limited by severe dental artifacts and image noise.

  9. Diagnostic accuracy of cone-beam CT in the assessment of mandibular invasion of lower gingival carcinoma: Comparison with conventional panoramic radiography

    International Nuclear Information System (INIS)

    Momin, Mohammad A.; Okochi, Kiyoshi; Watanabe, Hiroshi; Imaizumi, Akiko; Omura, Ken; Amagasa, Teruo; Okada, Norihiko; Ohbayashi, Naoto; Kurabayashi, Tohru

    2009-01-01

    Purpose: To evaluate the diagnostic accuracy of cone-beam CT in assessing mandibular invasion by lower gingival carcinoma and compare it with that of panoramic radiography. Patients and methods: Fifty patients with squamous cell carcinoma of the lower gingiva who were examined by both panoramic radiography and cone-beam CT before surgery were included in this study. Five radiologists used a 6-point rating scale to independently evaluate cone-beam CT and panoramic images for the presence or absence of alveolar bone and mandibular canal involvement by tumor. Using the histopathogical findings as the gold standard, we calculated and compared the area under the receiver operating characteristic curve (Az value) and the sensitivity and specificity of the two imaging modalities. Results: In evaluations of both alveolar bone and mandibular canal involvement, the mean Az value for cone-beam CT (0.918 and 0.977, respectively) was significantly higher than that for panoramic radiography (0.793 and 0.872, respectively). The mean sensitivity for cone-beam CT (89% and 99%, respectively) was significantly higher than that for panoramic radiography (73% and 56%, respectively). There was no significant difference in the mean specificity. While cone-beam CT could provide high-resolution three-dimensional images, the image quality around the alveolar crest was often hampered by severe dental artifacts and image noise, resulting in difficulties in detecting subtle alveolar invasion. Conclusion: Cone-beam CT was significantly superior to panoramic radiography in evaluating mandibular invasion by lower gingival carcinoma. Its diagnostic value in detecting subtle alveolar invasion, however, may be limited by severe dental artifacts and image noise.

  10. Should image rotation be addressed during routine cone-beam CT quality assurance?

    International Nuclear Information System (INIS)

    Ayan, Ahmet S; Lin Haibo; Yeager, Caitlyn; Deville, Curtiland; McDonough, James; Zhu, Timothy C; Anderson, Nathan; Ad, Voichita Bar; Both, Stefan; Lu, Hsiao-Ming

    2013-01-01

    The purpose of this study is to investigate whether quality assurance (QA) for cone-beam computed tomography (CBCT) image rotation is necessary in order to ensure the accuracy of CBCT based image-guided radiation therapy (IGRT) and adaptive radiotherapy (ART). Misregistration of angular coordinates during CBCT acquisition may lead to a rotated reconstructed image. If target localization is performed based on this image, an under- or over-dosage of the target volume (TV) and organs at risk (OARs) may occur. Therefore, patient CT image sets were rotated by 1° up to 3° and the treatment plans were recalculated to quantify changes in dose–volume histograms. A computer code in C++ was written to model the TV displacement and overlap area of an ellipse shape at the target and dose prescription levels corresponding to the image rotation. We investigated clinical scenarios in IGRT and ART in order to study the implications of image rotation on dose distributions for: (1) lateral TV and isocenter (SBRT), (2) central TV and isocenter (IMRT), (3) lateral TV and isocenter (IMRT). Mathematical analysis showed the dose coverage of TV depends on its shape, size, location, and orientation relative to the isocenter. Evaluation of three first scenario for θ = 1° showed variations in TV D95 in the context of IGRT and ART when compared to the original plan were within 2.7 ± 2.6% and 7.7 ± 6.9% respectively while variations in the second and third scenarios were less significant (<0.5%) for the angular range evaluated. However a larger degree of variation was found in terms of minimum and maximum doses for target and OARs. The rotation of CBCT image data sets may have significant dosimetric consequences in IGRT and ART. The TV's location relative to isocenter and shape determine the extent of alterations in dose indicators. Our findings suggest that a CBCT QA criterion of 1° would be a reasonable action level to ensure accurate dose delivery. (paper)

  11. Hypofractionated radiotherapy for lung tumors with online cone beam CT guidance and active breathing control

    Science.gov (United States)

    2010-01-01

    Background To study the set-up errors, PTV margin and toxicity of cone beam CT (CBCT) guided hypofractionated radiotherapy with active breathing control (ABC) for patients with non-small cell lung cancer (NSCLC) or metastatic tumors in lung. Methods 32 tumors in 20 patients were treated. Based on the location of tumor, dose per fraction given to tumor was divided into three groups: 12 Gy, 8 Gy and 6 Gy. ABC is applied for every patient. During each treatment, patients receive CBCT scan for online set-up correction. The pre- and post-correction setup errors between fractions, the interfractional and intrafractional, set-up errors, PTV margin as well as toxicity are analyzed. Results The pre-correction systematic and random errors in the left-right (LR), superior-inferior (SI), anterior-posterior (AP) directions were 3.7 mm and 5.3 mm, 3.1 mm and 2.1 mm, 3.7 mm and 2.8 mm, respectively, while the post-correction residual errors were 0.6 mm and 0.8 mm, 0.8 mm and 0.8 mm, 1.2 mm and 1.3 mm, respectively. There was an obvious intrafractional shift of tumor position. The pre-correction PTV margin was 9.5 mm in LR, 14.1 mm in SI and 8.2 mm in AP direction. After CBCT guided online correction, the PTV margin was markedly reduced in all three directions. The post-correction margins ranged 1.5 to 2.1 mm. The treatment was well tolerated by patients, of whom there were 4 (20%) grade1-2 acute pneumonitis, 3 (15%) grade1 acute esophagitis, 2 (10%) grade1 late pneumonitis and 1 (5%) grade 1 late esophagitis. Conclusion The positioning errors for lung SBRT using ABC were significant. Online correction with CBCT image guidance should be applied to reduce setup errors and PTV margin, which may reduce radiotherapy toxicity of tissues when ABC was used. PMID:20187962

  12. SU-E-I-59: Image Quality and Dose Measurement for Partial Cone-Beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Abouei, E; Ford, N [University of British Columbia, Vancouver, BC (Canada)

    2014-06-01

    Purpose: To characterize performance of cone beam CT (CBCT) used in dentistry investigating quantitatively the image quality and radiation dose during dental CBCT over different settings for partial rotation of the x-ray tube. Methods: Image quality and dose measurements were done on a variable field of view (FOV) dental CBCT (Carestream 9300). X-ray parameters for clinical settings were adjustable for 2–10 mA, 60–90 kVp, and two optional voxel size values, but time was fixed for each FOV. Image quality was assessed by scanning cylindrical poly-methyl methacrylate (PMMA) image quality phantom (SEDENTEXCT IQ), and then the images were analyzed using ImageJ to calculate image quality parameters such as noise, uniformity, and contrast to noise ratio (CNR). A protocol proposed by SEDENTEXCT, dose index 1 (DI1), was applied to dose measurements obtained using a thimble ionization chamber and cylindrical PMMA dose index phantom (SEDENTEXCT DI). Dose distributions were obtained using Gafchromic film. The phantoms were positioned in the FOV to imitate a clinical positioning. Results: The image noise was 6–12.5% which, when normalized to the difference of mean voxel value of PMMA and air, was comparable between different FOVs. Uniformity was 93.5ß 99.7% across the images. CNR was 1.7–4.2 and 6.3–14.3 for LDPE and Aluminum, respectively. Dose distributions were symmetric about the rotation angle's bisector. For large and medium FOVs at 4 mA and 80–90 kVp, DI1 values were in the range of 1.26–3.23 mGy. DI1 values were between 1.01–1.93 mGy for small FOV (5×5 cm{sup 2}) at 4–5 mA and 75–84 kVp. Conclusion: Noise decreased by increasing kVp, and the CNR increased for each FOV. When FOV size increased, image noise increased and CNR decreased. DI1 values were increased by increasing tube current (mA), tube voltage (kVp), and/or FOV. Funding for this project from NSERC Discovery grant, UBC Faculty of Dentistry Research Equipment Grant and UBC Faculty of

  13. Motion compensation in extremity cone-beam CT using a penalized image sharpness criterion

    Science.gov (United States)

    Sisniega, A.; Stayman, J. W.; Yorkston, J.; Siewerdsen, J. H.; Zbijewski, W.

    2017-05-01

    Cone-beam CT (CBCT) for musculoskeletal imaging would benefit from a method to reduce the effects of involuntary patient motion. In particular, the continuing improvement in spatial resolution of CBCT may enable tasks such as quantitative assessment of bone microarchitecture (0.1 mm-0.2 mm detail size), where even subtle, sub-mm motion blur might be detrimental. We propose a purely image based motion compensation method that requires no fiducials, tracking hardware or prior images. A statistical optimization algorithm (CMA-ES) is used to estimate a motion trajectory that optimizes an objective function consisting of an image sharpness criterion augmented by a regularization term that encourages smooth motion trajectories. The objective function is evaluated using a volume of interest (VOI, e.g. a single bone and surrounding area) where the motion can be assumed to be rigid. More complex motions can be addressed by using multiple VOIs. Gradient variance was found to be a suitable sharpness metric for this application. The performance of the compensation algorithm was evaluated in simulated and experimental CBCT data, and in a clinical dataset. Motion-induced artifacts and blurring were significantly reduced across a broad range of motion amplitudes, from 0.5 mm to 10 mm. Structure similarity index (SSIM) against a static volume was used in the simulation studies to quantify the performance of the motion compensation. In studies with translational motion, the SSIM improved from 0.86 before compensation to 0.97 after compensation for 0.5 mm motion, from 0.8 to 0.94 for 2 mm motion and from 0.52 to 0.87 for 10 mm motion (~70% increase). Similar reduction of artifacts was observed in a benchtop experiment with controlled translational motion of an anthropomorphic hand phantom, where SSIM (against a reconstruction of a static phantom) improved from 0.3 to 0.8 for 10 mm motion. Application to a clinical dataset of a lower extremity showed dramatic reduction

  14. Anatomical background and generalized detectability in tomosynthesis and cone-beam CT.

    Science.gov (United States)

    Gang, G J; Tward, D J; Lee, J; Siewerdsen, J H

    2010-05-01

    Anatomical background presents a major impediment to detectability in 2D radiography as well as 3D tomosynthesis and cone-beam CT (CBCT). This article incorporates theoretical and experimental analysis of anatomical background "noise" in cascaded systems analysis of 2D and 3D imaging performance to yield "generalized" metrics of noise-equivalent quanta (NEQ) and detectability index as a function of the orbital extent of the (circular arc) source-detector orbit. A physical phantom was designed based on principles of fractal self-similarity to exhibit power-law spectral density (kappa/Fbeta) comparable to various anatomical sites (e.g., breast and lung). Background power spectra [S(B)(F)] were computed as a function of source-detector orbital extent, including tomosynthesis (approximately 10 degrees -180 degrees) and CBCT (180 degrees + fan to 360 degrees) under two acquisition schemes: (1) Constant angular separation between projections (variable dose) and (2) constant total number of projections (constant dose). The resulting S(B) was incorporated in the generalized NEQ, and detectability index was computed from 3D cascaded systems analysis for a variety of imaging tasks. The phantom yielded power-law spectra within the expected spatial frequency range, quantifying the dependence of clutter magnitude (kappa) and correlation (beta) with increasing tomosynthesis angle. Incorporation of S(B) in the 3D NEQ provided a useful framework for analyzing the tradeoffs among anatomical, quantum, and electronic noise with dose and orbital extent. Distinct implications are posed for breast and chest tomosynthesis imaging system design-applications varying significantly in kappa and beta, and imaging task and, therefore, in optimal selection of orbital extent, number of projections, and dose. For example, low-frequency tasks (e.g., soft-tissue masses or nodules) tend to benefit from larger orbital extent and more fully 3D tomographic imaging, whereas high-frequency tasks (e

  15. Delayed-Phase Cone-Beam CT Improves Detectability of Intrahepatic Cholangiocarcinoma During Conventional Transarterial Chemoembolization

    Energy Technology Data Exchange (ETDEWEB)

    Schernthaner, Ruediger Egbert [The Johns Hopkins Hospital, Russell H. Morgan Department of Radiology and Radiological Science, Division of Vascular and Interventional Radiology (United States); Lin, MingDe [Philips Research North America, Ultrasound and Interventions (United States); Duran, Rafael; Chapiro, Julius; Wang, Zhijun; Geschwind, Jean-François, E-mail: jfg@jhmi.edu [The Johns Hopkins Hospital, Russell H. Morgan Department of Radiology and Radiological Science, Division of Vascular and Interventional Radiology (United States)

    2015-08-15

    PurposeTo evaluate the detectability of intrahepatic cholangiocarcinoma (ICC) on dual-phase cone-beam CT (DPCBCT) during conventional transarterial chemoembolization (cTACE) compared to that of digital subtraction angiography (DSA) with respect to pre-procedure contrast-enhanced magnetic resonance imaging (CE-MRI) of the liver.MethodsThis retrospective study included 17 consecutive patients (10 male, mean age 64) with ICC who underwent pre-procedure CE-MRI of the liver, and DSA and DPCBCT (early-arterial phase (EAP) and delayed-arterial phase (DAP)) just before cTACE. The visibility of each ICC lesion was graded by two radiologists on a three-rank scale (complete, partial, and none) on DPCBCT and DSA images, and then compared to pre-procedure CE-MRI.ResultsOf 61 ICC lesions, only 45.9 % were depicted by DSA, whereas EAP- and DAP-CBCT yielded a significantly higher detectability rate of 73.8 % and 93.4 %, respectively (p < 0.01). Out of the 33 lesions missed on DSA, 18 (54.5 %) and 30 (90.9 %) were revealed on EAP- and DAP-CBCT images, respectively. DSA depicted only one lesion that was missed by DPCBCT due to streak artifacts caused by a prosthetic mitral valve. DAP-CBCT identified significantly more lesions than EAP-CBCT (p < 0.01). Conversely, EAP-CBCT did not detect lesions missed by DAP-CBCT. For complete lesion visibility, DAP-CBCT yielded significantly higher detectability (78.7 %) compared to EAP (31.1 %) and DSA (21.3 %) (p < 0.01).ConclusionDPCBCT, and especially the DAP-CBCT, significantly improved the detectability of ICC lesions during cTACE compared to DSA. We recommend the routine use of DAP-CBCT in patients with ICC for per-procedure detectability and treatment planning in the setting of TACE.

  16. Task-based statistical image reconstruction for high-quality cone-beam CT

    Science.gov (United States)

    Dang, Hao; Webster Stayman, J.; Xu, Jennifer; Zbijewski, Wojciech; Sisniega, Alejandro; Mow, Michael; Wang, Xiaohui; Foos, David H.; Aygun, Nafi; Koliatsos, Vassilis E.; Siewerdsen, Jeffrey H.

    2017-11-01

    Task-based analysis of medical imaging performance underlies many ongoing efforts in the development of new imaging systems. In statistical image reconstruction, regularization is often formulated in terms to encourage smoothness and/or sharpness (e.g. a linear, quadratic, or Huber penalty) but without explicit formulation of the task. We propose an alternative regularization approach in which a spatially varying penalty is determined that maximizes task-based imaging performance at every location in a 3D image. We apply the method to model-based image reconstruction (MBIR—viz., penalized weighted least-squares, PWLS) in cone-beam CT (CBCT) of the head, focusing on the task of detecting a small, low-contrast intracranial hemorrhage (ICH), and we test the performance of the algorithm in the context of a recently developed CBCT prototype for point-of-care imaging of brain injury. Theoretical predictions of local spatial resolution and noise are computed via an optimization by which regularization (specifically, the quadratic penalty strength) is allowed to vary throughout the image to maximize local task-based detectability index ({{d}\\prime} ). Simulation studies and test-bench experiments were performed using an anthropomorphic head phantom. Three PWLS implementations were tested: conventional (constant) penalty; a certainty-based penalty derived to enforce constant point-spread function, PSF; and the task-based penalty derived to maximize local detectability at each location. Conventional (constant) regularization exhibited a fairly strong degree of spatial variation in {{d}\\prime} , and the certainty-based method achieved uniform PSF, but each exhibited a reduction in detectability compared to the task-based method, which improved detectability up to ~15%. The improvement was strongest in areas of high attenuation (skull base), where the conventional and certainty-based methods tended to over-smooth the data. The task-driven reconstruction method presents a

  17. Hypofractionated radiotherapy for lung tumors with online cone beam CT guidance and active breathing control

    International Nuclear Information System (INIS)

    Shen, Yali; Zhang, Hong; Wang, Jin; Zhong, Renming; Jiang, Xiaoqing; Xu, Qinfeng; Wang, Xin; Bai, Sen; Xu, Feng

    2010-01-01

    To study the set-up errors, PTV margin and toxicity of cone beam CT (CBCT) guided hypofractionated radiotherapy with active breathing control (ABC) for patients with non-small cell lung cancer (NSCLC) or metastatic tumors in lung. 32 tumors in 20 patients were treated. Based on the location of tumor, dose per fraction given to tumor was divided into three groups: 12 Gy, 8 Gy and 6 Gy. ABC is applied for every patient. During each treatment, patients receive CBCT scan for online set-up correction. The pre- and post-correction setup errors between fractions, the interfractional and intrafractional, set-up errors, PTV margin as well as toxicity are analyzed. The pre-correction systematic and random errors in the left-right (LR), superior-inferior (SI), anterior-posterior (AP) directions were 3.7 mm and 5.3 mm, 3.1 mm and 2.1 mm, 3.7 mm and 2.8 mm, respectively, while the post-correction residual errors were 0.6 mm and 0.8 mm, 0.8 mm and 0.8 mm, 1.2 mm and 1.3 mm, respectively. There was an obvious intrafractional shift of tumor position. The pre-correction PTV margin was 9.5 mm in LR, 14.1 mm in SI and 8.2 mm in AP direction. After CBCT guided online correction, the PTV margin was markedly reduced in all three directions. The post-correction margins ranged 1.5 to 2.1 mm. The treatment was well tolerated by patients, of whom there were 4 (20%) grade1-2 acute pneumonitis, 3 (15%) grade1 acute esophagitis, 2 (10%) grade1 late pneumonitis and 1 (5%) grade 1 late esophagitis. The positioning errors for lung SBRT using ABC were significant. Online correction with CBCT image guidance should be applied to reduce setup errors and PTV margin, which may reduce radiotherapy toxicity of tissues when ABC was used

  18. Hypofractionated radiotherapy for lung tumors with online cone beam CT guidance and active breathing control

    Directory of Open Access Journals (Sweden)

    Wang Xin

    2010-02-01

    Full Text Available Abstract Background To study the set-up errors, PTV margin and toxicity of cone beam CT (CBCT guided hypofractionated radiotherapy with active breathing control (ABC for patients with non-small cell lung cancer (NSCLC or metastatic tumors in lung. Methods 32 tumors in 20 patients were treated. Based on the location of tumor, dose per fraction given to tumor was divided into three groups: 12 Gy, 8 Gy and 6 Gy. ABC is applied for every patient. During each treatment, patients receive CBCT scan for online set-up correction. The pre- and post-correction setup errors between fractions, the interfractional and intrafractional, set-up errors, PTV margin as well as toxicity are analyzed. Results The pre-correction systematic and random errors in the left-right (LR, superior-inferior (SI, anterior-posterior (AP directions were 3.7 mm and 5.3 mm, 3.1 mm and 2.1 mm, 3.7 mm and 2.8 mm, respectively, while the post-correction residual errors were 0.6 mm and 0.8 mm, 0.8 mm and 0.8 mm, 1.2 mm and 1.3 mm, respectively. There was an obvious intrafractional shift of tumor position. The pre-correction PTV margin was 9.5 mm in LR, 14.1 mm in SI and 8.2 mm in AP direction. After CBCT guided online correction, the PTV margin was markedly reduced in all three directions. The post-correction margins ranged 1.5 to 2.1 mm. The treatment was well tolerated by patients, of whom there were 4 (20% grade1-2 acute pneumonitis, 3 (15% grade1 acute esophagitis, 2 (10% grade1 late pneumonitis and 1 (5% grade 1 late esophagitis. Conclusion The positioning errors for lung SBRT using ABC were significant. Online correction with CBCT image guidance should be applied to reduce setup errors and PTV margin, which may reduce radiotherapy toxicity of tissues when ABC was used.

  19. Automated classification of maxillofacial cysts in cone beam CT images using contourlet transformation and Spherical Harmonics.

    Science.gov (United States)

    Abdolali, Fatemeh; Zoroofi, Reza Aghaeizadeh; Otake, Yoshito; Sato, Yoshinobu

    2017-02-01

    Accurate detection of maxillofacial cysts is an essential step for diagnosis, monitoring and planning therapeutic intervention. Cysts can be of various sizes and shapes and existing detection methods lead to poor results. Customizing automatic detection systems to gain sufficient accuracy in clinical practice is highly challenging. For this purpose, integrating the engineering knowledge in efficient feature extraction is essential. This paper presents a novel framework for maxillofacial cysts detection. A hybrid methodology based on surface and texture information is introduced. The proposed approach consists of three main steps as follows: At first, each cystic lesion is segmented with high accuracy. Then, in the second and third steps, feature extraction and classification are performed. Contourlet and SPHARM coefficients are utilized as texture and shape features which are fed into the classifier. Two different classifiers are used in this study, i.e. support vector machine and sparse discriminant analysis. Generally SPHARM coefficients are estimated by the iterative residual fitting (IRF) algorithm which is based on stepwise regression method. In order to improve the accuracy of IRF estimation, a method based on extra orthogonalization is employed to reduce linear dependency. We have utilized a ground-truth dataset consisting of cone beam CT images of 96 patients, belonging to three maxillofacial cyst categories: radicular cyst, dentigerous cyst and keratocystic odontogenic tumor. Using orthogonalized SPHARM, residual sum of squares is decreased which leads to a more accurate estimation. Analysis of the results based on statistical measures such as specificity, sensitivity, positive predictive value and negative predictive value is reported. The classification rate of 96.48% is achieved using sparse discriminant analysis and orthogonalized SPHARM features. Classification accuracy at least improved by 8.94% with respect to conventional features. This study

  20. The role of off-focus radiation in scatter correction for dedicated cone beam breast CT.

    Science.gov (United States)

    Shi, Linxi; Vedantham, Srinivasan; Karellas, Andrew; Zhu, Lei

    2018-01-01

    Dedicated cone beam breast CT (CBBCT) suffers from x-ray scatter contamination. We aim to identify the source of the significant difference between the scatter distributions estimated by two recent methods proposed by our group and to investigate its effect on CBBCT image quality. We recently proposed two novel methods of scatter correction for CBBCT, using a library based (LB) technique and a forward projection (FP) model. Despite similar enhancement on CBBCT image qualities, these two methods obtain very different scatter distributions. We hypothesize that the off-focus radiation (OFR) is the contributor and results in nontrivial signals in x-ray projections, which is ignored in the scatter estimation via the LB method. Experiments using a thin wire test tool are designed to study the effect of OFR on CBBCT spatial resolution by measuring the point spread function (PSF) and the modulation transfer function (MTF). A narrow collimator setting is used to suppress the OFR-induced signals. In addition, "PSFs" and "MTFs" are measured on clinical CBBCT images obtained by the LB and FP methods using small calcifications as point sources. The improvement of spatial resolution achieved by suppressing OFR in the wire experiment as well as in the clinical study is quantified by the improvement ratios of PSFs and spatial frequencies at different MTF values. Our hypothesis that OFR causes the imaging difference between the FP and LB methods is verified if these ratios obtained from experimental and clinical data are consistent. In the wire experiment, the results show that suppression of OFR increases the maximum signal of the PSF by about 14% and reduces the full-width-at-half-maximum (FWHM) by about 12.0%. Similar improvement on spatial resolution is achieved by the FP method compared with the LB method in the patient study. The improvement ratios of spatial frequencies at different MTF values without OFR match very well in both studies at a level of around 16%, with an

  1. Low-Dose and Scatter-Free Cone-Beam CT Imaging Using a Stationary Beam Blocker in a Single Scan: Phantom Studies

    Directory of Open Access Journals (Sweden)

    Xue Dong

    2013-01-01

    Full Text Available Excessive imaging dose from repeated scans and poor image quality mainly due to scatter contamination are the two bottlenecks of cone-beam CT (CBCT imaging. Compressed sensing (CS reconstruction algorithms show promises in recovering faithful signals from low-dose projection data but do not serve well the needs of accurate CBCT imaging if effective scatter correction is not in place. Scatter can be accurately measured and removed using measurement-based methods. However, these approaches are considered unpractical in the conventional FDK reconstruction, due to the inevitable primary loss for scatter measurement. We combine measurement-based scatter correction and CS-based iterative reconstruction to generate scatter-free images from low-dose projections. We distribute blocked areas on the detector where primary signals are considered redundant in a full scan. Scatter distribution is estimated by interpolating/extrapolating measured scatter samples inside blocked areas. CS-based iterative reconstruction is finally carried out on the undersampled data to obtain scatter-free and low-dose CBCT images. With only 25% of conventional full-scan dose, our method reduces the average CT number error from 250 HU to 24 HU and increases the contrast by a factor of 2.1 on Catphan 600 phantom. On an anthropomorphic head phantom, the average CT number error is reduced from 224 HU to 10 HU in the central uniform area.

  2. Cone-beam CT image contrast and attenuation-map linearity improvement (CALI) for brain stereotactic radiosurgery procedures

    Science.gov (United States)

    Hashemi, Sayed Masoud; Lee, Young; Eriksson, Markus; Nordström, Hâkan; Mainprize, James; Grouza, Vladimir; Huynh, Christopher; Sahgal, Arjun; Song, William Y.; Ruschin, Mark

    2017-03-01

    A Contrast and Attenuation-map (CT-number) Linearity Improvement (CALI) framework is proposed for cone-beam CT (CBCT) images used for brain stereotactic radiosurgery (SRS). The proposed framework is used together with our high spatial resolution iterative reconstruction algorithm and is tailored for the Leksell Gamma Knife ICON (Elekta, Stockholm, Sweden). The incorporated CBCT system in ICON facilitates frameless SRS planning and treatment delivery. The ICON employs a half-cone geometry to accommodate the existing treatment couch. This geometry increases the amount of artifacts and together with other physical imperfections causes image inhomogeneity and contrast reduction. Our proposed framework includes a preprocessing step, involving a shading and beam-hardening artifact correction, and a post-processing step to correct the dome/capping artifact caused by the spatial variations in x-ray energy generated by bowtie-filter. Our shading correction algorithm relies solely on the acquired projection images (i.e. no prior information required) and utilizes filtered-back-projection (FBP) reconstructed images to generate a segmented bone and soft-tissue map. Ideal projections are estimated from the segmented images and a smoothed version of the difference between the ideal and measured projections is used in correction. The proposed beam-hardening and dome artifact corrections are segmentation free. The CALI was tested on CatPhan, as well as patient images acquired on the ICON system. The resulting clinical brain images show substantial improvements in soft contrast visibility, revealing structures such as ventricles and lesions which were otherwise un-detectable in FBP-reconstructed images. The linearity of the reconstructed attenuation-map was also improved, resulting in more accurate CT#.

  3. MO-F-BRA-02: Evaluation of 4D CT to 4D Cone-Beam CT Deformable Image Registration for Lung Cancer Adaptive Radiation Therapy.

    Science.gov (United States)

    Balik, S; Hugo, G; Weiss, E; Jan, N; Roman, N; Sleeman, W; Fatyga, M; Christensen, G; Murphy, M; Lu, J; Keall, P; Williamson, J

    2012-06-01

    To evaluate two deformable image registration (DIR) algorithms for the purpose of contour mapping to support image guided adaptive radiotherapy (IGART) with 4D cone beam CT (4DCBCT). Eleven locally advanced non-small cell lung cancer (NSCLC) patients underwent one planning 4D fan- beam CT (4DFBCT) and seven weekly 4DCBCT scans. Gross tumor volume (GTV) and carina were delineated by a physician in all 4D images. For day to day registration, the end of inspiration 4DFBCT phase was deformably registered to the corresponding phase in each 4DCBCT image. For phase to phase registration, the end of inspiration phase from each 4D image was registered to end of expiration phase. The delineated contours were warped using the resulting transforms and compared to the manual contours through Dice similarity coefficient (DSC), false positive and false negative indices, and, for carina, target registration error (TRE). Two DIR algorithms were tested: 1) small deformation, inverse consistent linear elastic (SICLE) algorithm and 2) Insight Toolkit diffeomorphic demons (DEMONS). For day to day registrations, the mean DSC was 0.59 ± 0.16 after rigid registration, 0.72 ± 0.13 with SICLE and to 0.66 ± 0.18 with DEMONS. SICLE and DEMONS reduced TRE to 4.1 ± 2.1 mm and 5.8 ± 3.7 mm respectively, from 6.2 ± 3.5 mm; and reduced false positive index to 0.27 and 0.26 respectively from 0.46. Registration with the cone beam as the fixed image resulted in higher DSC than with the fan beam as fixed (p < 0.001). SICLE and DEMONS increased the DSC on average by 10.0% and 8.0% and reduced TRE by 2.8 mm and 2.9 mm respectively for phase to phase DIR. DIR achieved more congruent mapping of target structures to delineations than rigid registration alone, although DIR performance varied with algorithm and patient. This work was supported by National Cancer Institute Grant No. P01 CA 116602. © 2012 American Association of Physicists in Medicine.

  4. High-resolution helical cone-beam micro-CT with theoretically-exact reconstruction from experimental data

    International Nuclear Information System (INIS)

    Varslot, T.; Kingston, A.; Myers, G.; Sheppard, A.

    2011-01-01

    conventional circular scan micro-CT. Conclusions: Autofocus-corrected, theoretically-exact cone-beam reconstruction is a viable option for reducing acquisition time in high-resolution micro-CT imaging. It also opens up the possibility of efficiently imaging long objects.

  5. A new method to determine the projected coordinate origin of a cone-beam CT system using elliptical projection

    International Nuclear Information System (INIS)

    Yang Min; Li Baolei; Jin Xuling

    2010-01-01

    In order to determine the projected coordinate origin in the cone-beam CT scanning system with respect to the Feldkamp-Davis-Kress (FDK) algorithm,we propose a simple yet feasible method to accurately measure the projected coordinate origin. This method was established on the basis of the theory that the projection of a spherical object in the cone-beam field is an ellipse. We first utilized image processing and the least square estimation method to get each major axis of the elliptical Digital Radiography (DR) projections of a group of spherical objects. Then we determined the intersection point of the group of major axis by solving an over-determined equation set that was composed by the major axis equations of all the elliptical projections. Based on the experimental results,this new method was proved to be easy to implement in practical scanning systems with high accuracy and anti-noise capability. (authors)

  6. Development of the three dimensional image display program for limited cone beam X-ray CT for dental use (Ortho-CT)

    International Nuclear Information System (INIS)

    Arai, Yoshinori; Hashimoto, Koji; Shinoda, Koji

    2000-01-01

    We have already developed and reported a limited cone beam X-ray CT system for dental use (Ortho-CT). This system has been used clinically since 1997. In this study, we report a 3D surface display program for Ortho-CT which has been newly-developed by the authors. The 3D surface display software has been developed using visual C ++ (Microsoft Co. WA. USA) and a personal computer (Pentium 450MHz Intel Co. CA USA, Windows NT 4.0 Microsoft WA. USA). In this software, the 3D surface images are recorded as AVI files and can be displayed on the personal computer. The 3D images can be rotated and a stepwise change of the threshold voxel value for binary image formation can be automatically used. We have applied these 3D surface images to clinical studies from January 1999 to May 1999 at the Radiology section in our Dental hospital. The images can be displayed very easily in personal computers using AVI files. Thirty-five cases have been reconstructed using 3D surface images in this way. The 3D surface image is useful in the diagnosis of fractures of the mandibular head and impacted teeth. Only teeth are observed when a relative threshold voxel value is set at a high level such as about 0.37. When the threshold is changed to a lower value (about 0.3), we can observe both teeth and the surface of the bone. We have developed a 3D surface display program for personal computers. The images are useful for the diagnosis of the pathosis in the maxillofacial region. (author)

  7. SU-E-J-113: The Influence of Optimizing Pediatric CT Simulator Protocols On the Treatment Dose Calculation in Radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Zhang, Y; Zhang, J; Hu, Q; Tie, J; Wu, H [Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education), Department of Radiotherapy, Peking University Cancer Hospital ' Institute, Beijing (China); Deng, J [Department of Therapeutic Radiology, Yale University, New Haven, CT (United States)

    2014-06-01

    Purpose: To investigate the possibility of applying optimized scanning protocols for pediatric CT simulation by quantifying the dosimetric inaccuracy introduced by using a fixed HU to density conversion. Methods: The images of a CIRS electron density reference phantom (Model 062) were acquired by a Siemens CT simulator (Sensation Open) using the following settings of tube voltage and beam current: 120 kV/190mA (the reference protocol used to calibrate CT for our treatment planning system (TPS)); Fixed 190mA combined with all available kV: 80, 100, and 140; fixed 120 kV and various current from 37 to 444 mA (scanner extremes) with interval of 30 mA. To avoid the HU uncertainty of point sampling in the various inserts of known electron densities, the mean CT numbers of the central cylindrical volume were calculated using DICOMan software. The doses per 100 MU to the reference point (SAD=100cm, Depth=10cm, Field=10X10cm, 6MV photon beam) in a virtual cubic phantom (30X30X30cm) were calculated using Eclipse TPS (calculation model: AcurosXB-11031) by assigning the CT numbers to HU of typical materials acquired by various protocols. Results: For the inserts of densities less than muscle, CT number fluctuations of all protocols were within the tolerance of 10 HU as accepted by AAPM-TG66. For more condensed materials, fixed kV yielded stable HU with any mA combination where largest disparities were found in 1750mg/cc insert: HU{sub reference}=1801(106.6cGy), HU{sub minimum}=1799 (106.6cGy, error{sub dose}=0.00%), HU{sub maximum}=1815 (106.8cGy, error{sub dose}=0.19%). Yet greater disagreements were observed with increasing density when kV was modified: HU{sub minimum}=1646 (104.5cGy, error{sub dose}=- 1.97%), HU{sub maximum}=2487 (116.4cGy, error{sub dose}=9.19%) in 1750mg/cc insert. Conclusion: Without affecting treatment dose calculation, personalized mA optimization of CT simulator can be conducted by fixing kV for a better cost-effectiveness of imaging dose and quality

  8. Assessment of contrast enhanced respiration managed cone-beam CT for image guided radiotherapy of intrahepatic tumors

    Energy Technology Data Exchange (ETDEWEB)

    Jensen, Nikolaj K. G., E-mail: nkyj@regionsjaelland.dk [Physics and Engineering, London Regional Cancer Program, London, Ontario N6A3K7 (Canada); Stewart, Errol [Radiology, St. Joseph' s Health Care, London, Ontario N6A 4V2 (Canada); Imaging Research Lab, Robarts Research Institute, London, Ontario N6A 5B7 (Canada); Imaging Program, Lawson Health Research Institute, London, Ontario N6C 2R5 (Canada); Lock, Michael; Fisher, Barbara [Radiation Oncology, London Regional Cancer Program, London, Ontario N6A3K7 (Canada); Department of Oncology, University of Western Ontario, London, Ontario N6A 4L6 (Canada); Kozak, Roman [Radiology, St. Joseph' s Health Care, London, Ontario N6A 4V2 (Canada); Chen, Jeff [Physics and Engineering, London Regional Cancer Program, London, Ontario N6A3K7 (Canada); Department of Oncology, University of Western Ontario, London, Ontario N6A 4L6 (Canada); Department of Medical Biophysics, University of Western Ontario, London, Ontario N6A 5C1 (Canada); Lee, Ting-Yim [Radiology, St. Joseph' s Health Care, London, Ontario N6A 4V2 (Canada); Imaging Research Lab, Robarts Research Institute, London, Ontario N6A 5B7 (Canada); Imaging Program, Lawson Health Research Institute, London, Ontario N6C 2R5 (Canada); Department of Oncology, University of Western Ontario, London, Ontario N6A 4L6 (Canada); Department of Medical Biophysics, University of Western Ontario, London, Ontario N6A 5C1 (Canada); Wong, Eugene [Physics and Engineering, London Regional Cancer Program, London, Ontario N6A3K7 (Canada); Department of Oncology, University of Western Ontario, London, Ontario N6A 4L6 (Canada); Department of Medical Biophysics, University of Western Ontario, London, Ontario N6A 5C1 (Canada); Department of Physics and Astronomy, University of Western Ontario, London, Ontario N6A 3K7 (Canada)

    2014-05-15

    Purpose: Contrast enhancement and respiration management are widely used during image acquisition for radiotherapy treatment planning of liver tumors along with respiration management at the treatment unit. However, neither respiration management nor intravenous contrast is commonly used during cone-beam CT (CBCT) image acquisition for alignment prior to radiotherapy. In this study, the authors investigate the potential gains of injecting an iodinated contrast agent in combination with respiration management during CBCT acquisition for liver tumor radiotherapy. Methods: Five rabbits with implanted liver tumors were subjected to CBCT with and without motion management and contrast injection. The acquired CBCT images were registered to the planning CT to determine alignment accuracy and dosimetric impact. The authors developed a simulation tool for simulating contrast-enhanced CBCT images from dynamic contrast enhanced CT imaging (DCE-CT) to determine optimal contrast injection protocols. The tool was validated against contrast-enhanced CBCT of the rabbit subjects and was used for five human patients diagnosed with hepatocellular carcinoma. Results: In the rabbit experiment, when neither motion management nor contrast was used, tumor centroid misalignment between planning image and CBCT was 9.2 mm. This was reduced to 2.8 mm when both techniques were employed. Tumors were not visualized in clinical CBCT images of human subjects. Simulated contrast-enhanced CBCT was found to improve tumor contrast in all subjects. Different patients were found to require different contrast injections to maximize tumor contrast. Conclusions: Based on the authors’ animal study, respiration managed contrast enhanced CBCT improves IGRT significantly. Contrast enhanced CBCT benefits from patient specific tracer kinetics determined from DCE-CT.

  9. Assessment of contrast enhanced respiration managed cone-beam CT for image guided radiotherapy of intrahepatic tumors

    International Nuclear Information System (INIS)

    Jensen, Nikolaj K. G.; Stewart, Errol; Lock, Michael; Fisher, Barbara; Kozak, Roman; Chen, Jeff; Lee, Ting-Yim; Wong, Eugene

    2014-01-01

    Purpose: Contrast enhancement and respiration management are widely used during image acquisition for radiotherapy treatment planning of liver tumors along with respiration management at the treatment unit. However, neither respiration management nor intravenous contrast is commonly used during cone-beam CT (CBCT) image acquisition for alignment prior to radiotherapy. In this study, the authors investigate the potential gains of injecting an iodinated contrast agent in combination with respiration management during CBCT acquisition for liver tumor radiotherapy. Methods: Five rabbits with implanted liver tumors were subjected to CBCT with and without motion management and contrast injection. The acquired CBCT images were registered to the planning CT to determine alignment accuracy and dosimetric impact. The authors developed a simulation tool for simulating contrast-enhanced CBCT images from dynamic contrast enhanced CT imaging (DCE-CT) to determine optimal contrast injection protocols. The tool was validated against contrast-enhanced CBCT of the rabbit subjects and was used for five human patients diagnosed with hepatocellular carcinoma. Results: In the rabbit experiment, when neither motion management nor contrast was used, tumor centroid misalignment between planning image and CBCT was 9.2 mm. This was reduced to 2.8 mm when both techniques were employed. Tumors were not visualized in clinical CBCT images of human subjects. Simulated contrast-enhanced CBCT was found to improve tumor contrast in all subjects. Different patients were found to require different contrast injections to maximize tumor contrast. Conclusions: Based on the authors’ animal study, respiration managed contrast enhanced CBCT improves IGRT significantly. Contrast enhanced CBCT benefits from patient specific tracer kinetics determined from DCE-CT

  10. Dosimetric variation due to CT inter-slice spacing in four-dimensional carbon beam lung therapy

    International Nuclear Information System (INIS)

    Kumagai, Motoki; Mori, Shinichiro; Kandatsu, Susumu; Baba, Masayuki; Sharp, Gregory C; Asakura, Hiroshi; Endo, Masahiro

    2009-01-01

    When CT data with thick slice thickness are used in treatment planning, geometrical uncertainty may induce dosimetric errors. We evaluated carbon ion dose variations due to different CT slice thicknesses using a four-dimensional (4D) carbon ion beam dose calculation, and compared results between ungated and gated respiratory strategies. Seven lung patients were scanned in 4D mode with a 0.5 mm slice thickness using a 256-multi-slice CT scanner. CT images were averaged with various numbers of images to simulate reconstructed images with various slice thicknesses (0.5-5.0 mm). Two scenarios were studied (respiratory-ungated and -gated strategies). Range compensators were designed for each of the CT volumes with coarse inter-slice spacing to cover the internal target volume (ITV), as defined from 4DCT. Carbon ion dose distribution was computed for each resulting ITV on the 0.5 mm slice 4DCT data. The accumulated dose distribution was then calculated using deformable registration for 4D dose assessment. The magnitude of over- and under-dosage was found to be larger with the use of range compensators designed with a coarser inter-slice spacing than those obtained with a 0.5 mm slice thickness. Although no under-dosage was observed within the clinical target volume (CTV) region, D95 remained at over 97% of the prescribed dose for the ungated strategy and 95% for the gated strategy for all slice thicknesses. An inter-slice spacing of less than 3 mm may be able to minimize dose variation between the ungated and gated strategies. Although volumes with increased inter-slice spacing may reduce geometrical accuracy at a certain respiratory phase, this does not significantly affect delivery of the accumulated dose to the target during the treatment course.

  11. Automated double-cone-beam CT fusion technique. Enhanced evaluation of glue distribution in cases of spinal dural arteriovenous fistula (SDAVF) embolisation.

    Science.gov (United States)

    Faragò, Giuseppe; Caldiera, V; Antozzi, C; Bellino, A; Innocenti, A; Ciceri, E

    2017-05-01

    Spinal dural arteriovenous fistulas (SDAVFs) are acquired diseases that represent the majority of all arteriovenous spinal shunts, leading to progressive and disabling myelopathy. Treatment is focused on accurately disconnecting the fistula point. We present our experience with the double-cone-beam CT fusion technique successfully applied to evaluate treatment results in a series of SDAVFs. Between November 2011 and December 2015 we performed double-DynaCT acquisition (pre- and post-embolisation) in 12 cases of SDAVF. A successful DynaCT fusion technique was only achieved in the group of patients with pre- and post-treatment images acquired at the same time as the treatment session, under general anaesthesia (4/12). DynaCT performed on different days proved to be inadequate for the automated fusion technique because of changes in the body position (8/12). A pre-treatment flat-panel cone-beam CT with contrast, at the time of diagnostic angiography, can be very helpful to detect the correct level of the fistula and the relationship between the fistula and the surrounding structures. In case of the endovascular approach, additional post-treatment native acquisition merged with the pre-treatment acquisition (double-cone-beam CT fusion technique) permits to immediately evaluate the distribution of the glue cast and to confirm the success of the procedure. • SDAVF treatment must be aimed to occlude the fistula point shunt. • Native post-operative cone-beam CT permits high-spatial-resolution imaging of the embolic cast. • The automated double-cone-beam CT fusion technique (pre/post) accurately demonstrates intravascular glue distribution after embolisation. • Patient movements should be avoided to obtain good technical results.

  12. kV x-ray dual digital tomosynthesis for image guided lung SBRT

    Science.gov (United States)

    Partain, Larry; Boyd, Douglas; Kim, Namho; Hernandez, Andrew; Daly, Megan; Boone, John

    2016-03-01

    Two simulated sets of digital tomosynthesis images of the lungs, each acquired at a 90 degree angle from the other, with 19 projection images used for each set and SART iterative reconstructed, gives dual tomosynthesis slice image quality approaching that of spiral CT, and with a data acquisition time that is 3% of that of cone beam CT. This fast kV acquisition, should allow near real time tracking of lung tumors in patients receiving SBRT, based on a novel TumoTrakTM multi-source X-ray tube design. Until this TumoTrakTM prototype is completed over the next year, its projected performance was simulated from the DRR images created from a spiral CT data set from a lung cancer patient. The resulting dual digital tomosynthesis reconstructed images of the lung tumor were exceptional and approached that of the gold standard Feldkamp CT reconstruction of breath hold, diagnostic, spiral, multirow, CT data. The relative dose at 46 mAs was less than 10% of what it would have been if the digital tomosynthesis had been done at the 472 mAs of the CT data set. This is for a 0.77 fps imaging rate sufficient to resolve respiratory motion in many free breathing patients during SBRT. Such image guidance could decrease the magnitudes of targeting error margins by as much as 20 mm or more in the craniocaudal direction for lower lobe lesions while markedly reducing dose to normal lung, heart and other critical structures. These initial results suggest a wide range of topics for future work.

  13. Gambaran densitas kamar pulpa gigi sulung menggunakan cone beam CT-3D (Description of pulp chamber density in deciduous teeth using cone beam CT-3D

    Directory of Open Access Journals (Sweden)

    Herdiyati Y

    2013-06-01

    Full Text Available Background: Dental caries is the most common chronic diseases. Detection of caries is needed, especially on the deciduous teeth. An examination such as radiological examination is essential. The radiographic figures distinguish radiolucent of the crown. Digital radiography cone beam computed tomography (CBCT is able to show a more detailed picture. Purpose: This study was aimed to get value of the density of pulp chamber of caries and non caries deciduous teeth using CBCT radiographs. Methods: The study was conducted by using simple descriptive. The samples were all the data CBCT of pediatric patients aged 7-10 years who visited the Dental Hospital of the Faculty of Dentistry, University of Padjadjaran. The samples were teeth with single and double root. Results: The results showed that the value of the normal pulp density is 422.56 Hu, while the condition of caries decreased becomes -77.89 Hu. Conclusion: The tooth with caries showed a lower density than the non caries/tooth.Latar belakang: Karies gigi merupakan penyakit kronis yang sering terjadi. Deteksi terhadap karies sangat diperlukan terutama pada gigi decidius. Pemeriksaan penunjang berupa pemeriksaan radiologis sangat diperlukan. Secara umum gambaran radiografi dapat membedakan karies berupa gambaran radiolusent pada mahkota. Radiografi digital cone beam computed tomografi (CBCT, merupakan jenis radiografi yang mampu memperlihatkan gambaran yang lebih detail. Tujuan: Penelitian ini bertujuan mendapatkan nilai densitas kamar pulpa gigi sulung yang karies dan non karies menggunakan radiografi CBCT. Metode: Penelitian dilakukan dengan metode simple deskriptif. Sampel penelitian adalah semua data CBCT dari pasien anak berusia 7 - 10 tahun yang berkunjung ke RSGM Fakultas Kedokteran Gigi Universitas Padjadjaran. Gigi yang dianalisa meliputi gigi berakar tunggal dan berakar ganda. Hasil: Hasil penelitian menunjukkan bahwa nilai densitas pulpa normal adalah 422,56 Hu, sedangkan pada kondisi

  14. An Approximate Cone Beam Reconstruction Algorithm for Gantry-Tilted CT Using Tangential Filtering

    Directory of Open Access Journals (Sweden)

    Ming Yan

    2006-01-01

    Full Text Available FDK algorithm is a well-known 3D (three-dimensional approximate algorithm for CT (computed tomography image reconstruction and is also known to suffer from considerable artifacts when the scanning cone angle is large. Recently, it has been improved by performing the ramp filtering along the tangential direction of the X-ray source helix for dealing with the large cone angle problem. In this paper, we present an FDK-type approximate reconstruction algorithm for gantry-tilted CT imaging. The proposed method improves the image reconstruction by filtering the projection data along a proper direction which is determined by CT parameters and gantry-tilted angle. As a result, the proposed algorithm for gantry-tilted CT reconstruction can provide more scanning flexibilities in clinical CT scanning and is efficient in computation. The performance of the proposed algorithm is evaluated with turbell clock phantom and thorax phantom and compared with FDK algorithm and a popular 2D (two-dimensional approximate algorithm. The results show that the proposed algorithm can achieve better image quality for gantry-tilted CT image reconstruction.

  15. Auxiliary KCNE subunits modulate both homotetrameric Kv2.1 and heterotetrameric Kv2.1/Kv6.4 channels

    DEFF Research Database (Denmark)

    David, Jens-Peter; Stas, Jeroen I.; Schmitt, Nicole

    2015-01-01

    The diversity of the voltage-gated K+ (Kv) channel subfamily Kv2 is increased by interactions with auxiliary β-subunits and by assembly with members of the modulatory so-called silent Kv subfamilies (Kv5-Kv6 and Kv8-Kv9). However, it has not yet been investigated whether these two types of modula...

  16. Diagnostic accuracy of the detection of bone change using panoramic TMJ projection. Comparative study with limited cone-beam CT

    International Nuclear Information System (INIS)

    Araki, Kazuyuki; Okano, Tomohiro; Kataoka, Ryuta; Honda, Kazuya; Endo, Atsushi; Kaneko, Norikazu; Funahashi, Itsuo

    2008-01-01

    Panoramic temporoman joint (TMJ) projection is one of the alternative methods of conventional radiography, such as transcranial projection, for diagnosing temporomandibular joint disorder. There have been a few reports describing the diagnostic ability of this method. We evaluated the diagnostic accuracy of detecting bone change with panoramic TMJ projection. Fifty TMJs in 25 patients were examined. All TMJs were examined by panoramic TMJ projection (Hyper XF) and limited cone-beam CT (3D Accuitomo FPD; 3DX). Two observers evaluated the presence of bone change in the TMJ region using panoramic TMJ projection. One other observer evaluated the limited cone-beam CT for the presence and the pattern of bone changes in the TMJ region as the gold standard. Panoramic TMJ findings were evaluated with regard to sensitivity, specificity, and accuracy. Sensitivity, specificity and accuracy of the panoramic TMJ projection were 0.86, 0.76, and 0.82, respectively. These results and those of previous reports on other radiographic methods for TMJ suggest that panoramic TMJ projection is a useful method of screening for bone change due to TMJ disorder. (author)

  17. Three-dimensional image analysis of a head of the giant panda by the cone-beam type CT

    International Nuclear Information System (INIS)

    Endo, H.; Komiya, T.; Narushima, E.; Suzuki, N.

    2002-01-01

    The cone-beam type CT (Computed Tomography) enabled us to collect the three-dimensional (3D) digitalized data directly from the animal carcass. In this study, we applied the techniques of the cone-beam type CT for a carcass head of the giant panda (Ailuropoda melanoleuca) to obtain the 3D images easily without reconstruction process, and could morphologically examine the sections from the 3D data by means of non-destructive observations. The important results of the study represent the two following points. 1) We could show the morphological relationships between the muscles of mastication and the mandible in non-destructive status from the 3D data. The exact position of the coronoid process could be recognized in the rostro-lateral space of the temporal fossa. 2) By the serial sections from the 3D data sets, the morphological characteristics in the nasal cavity were detailed with high resolution in this rare species. The nasal concha was well-developed in the nasal cavity. The ethmoidal labyrinth was encountered immediately caudal to the nasal cavity and close to the region of the olfactory bulb. The ethmoidal labyrinth consisted of the complicated osseous structure in this area. The data will be useful to discuss the olfactory function in the reproduction behavior of this species

  18. A proposed method for accurate 3D analysis of cochlear implant migration using fusion of cone beam CT

    Directory of Open Access Journals (Sweden)

    Guido eDees

    2016-01-01

    Full Text Available IntroductionThe goal of this investigation was to compare fusion of sequential cone beam CT volumes to the gold standard (fiducial registration in order to be able to analyze clinical CI migration with high accuracy in three dimensions. Materials and MethodsPaired time-lapsed cone beam CT volumes were performed on five human cadaver temporal bones and one human subject. These volumes were fused using 3D Slicer 4 and BRAINSFit software. Using a gold standard fiducial technique, the accuracy, robustness and performance time of the fusion process were assessed.Results This proposed fusion protocol achieves a sub voxel mean Euclidean distance of 0.05 millimeter in human cadaver temporal bones and 0.16 millimeter when applied to the described in vivo human synthetic data set in over 95% of all fusions. Performance times are less than two minutes.ConclusionHere a new and validated method based on existing techniques is described which could be used to accurately quantify migration of cochlear implant electrodes.

  19. Cone beam CT imaging with limited angle of projections and prior knowledge for volumetric verification of non-coplanar beam radiation therapy: a proof of concept study

    International Nuclear Information System (INIS)

    Meng, Bowen; Xing, Lei; Han, Bin; Koong, Albert; Chang, Daniel; Li, Ruijiang; Cheng, Jason

    2013-01-01

    Non-coplanar beams are important for treatment of both cranial and noncranial tumors. Treatment verification of such beams with couch rotation/kicks, however, is challenging, particularly for the application of cone beam CT (CBCT). In this situation, only limited and unconventional imaging angles are feasible to avoid collision between the gantry, couch, patient, and on-board imaging system. The purpose of this work is to develop a CBCT verification strategy for patients undergoing non-coplanar radiation therapy. We propose an image reconstruction scheme that integrates a prior image constrained compressed sensing (PICCS) technique with image registration. Planning CT or CBCT acquired at the neutral position is rotated and translated according to the nominal couch rotation/translation to serve as the initial prior image. Here, the nominal couch movement is chosen to have a rotational error of 5° and translational error of 8 mm from the ground truth in one or more axes or directions. The proposed reconstruction scheme alternates between two major steps. First, an image is reconstructed using the PICCS technique implemented with total-variation minimization and simultaneous algebraic reconstruction. Second, the rotational/translational setup errors are corrected and the prior image is updated by applying rigid image registration between the reconstructed image and the previous prior image. The PICCS algorithm and rigid image registration are alternated iteratively until the registration results fall below a predetermined threshold. The proposed reconstruction algorithm is evaluated with an anthropomorphic digital phantom and physical head phantom. The proposed algorithm provides useful volumetric images for patient setup using projections with an angular range as small as 60°. It reduced the translational setup errors from 8 mm to generally <1 mm and the rotational setup errors from 5° to <1°. Compared with the PICCS algorithm alone, the integration of rigid

  20. Dose cone-beam CT alter treatment plans? Comparison of preoperative implant planning using panoramic versus cone-beam CT images

    Energy Technology Data Exchange (ETDEWEB)

    Guerrero, Maria Eugenia; Jacobs, Reinhilde [Dept. of Oral and Maxillofacial Surgery, University Hospitals, Leuven (Belgium); Norge, Jorge; Castro, Carmen [Master of Periodontology, Universidad San Martin de Porres, Lima (Peru)

    2014-06-15

    The present study was performed to compare the planning of implant placement based on panoramic radiography (PAN) and cone-beam computed tomography (CBCT) images, and to study the impact of the image dataset on the treatment planning. One hundred five partially edentulous patients (77 males, 28 females, mean age: 46 years, range: 26-67 years) seeking oral implant rehabilitation were referred for presurgical imaging. Imaging consisted of PAN and CBCT imaging. Four observers planned implant treatment based on the two-dimensional (2D) image datasets and at least one month later on the three-dimensional (3D) image dataset. Apart from presurgical diagnostic and dimensional measurement tasks, the observers needed to indicate the surgical confidence levels and assess the image quality in relation to the presurgical needs. All observers confirmed that both imaging modalities (PAN and CBCT) gave similar values when planning implant diameter. Also, the results showed no differences between both imaging modalities for the length of implants with an anterior location. However, significant differences were found in the length of implants with a posterior location. For implant dimensions, longer lengths of the implants were planned with PAN, as confirmed by two observers. CBCT provided images with improved scores for subjective image quality and surgical confidence levels. Within the limitations of this study, there was a trend toward PAN-based preoperative planning of implant placement leading towards the use of longer implants within the posterior jaw bone.

  1. Intraoperative imaging for patient safety and QA: detection of intracranial hemorrhage using C-arm cone-beam CT

    Science.gov (United States)

    Schafer, Sebastian; Wang, Adam; Otake, Yoshito; Stayman, J. W.; Zbijewski, Wojciech; Kleinszig, Gerhard; Xia, Xuewei; Gallia, Gary L.; Siewerdsen, Jeffrey H.

    2013-03-01

    Intraoperative imaging could improve patient safety and quality assurance (QA) via the detection of subtle complications that might otherwise only be found hours after surgery. Such capability could therefore reduce morbidity and the need for additional intervention. Among the severe adverse events that could be more quickly detected by high-quality intraoperative imaging is acute intracranial hemorrhage (ICH), conventionally assessed using post-operative CT. A mobile C-arm capable of high-quality cone-beam CT (CBCT) in combination with advanced image reconstruction techniques is reported as a means of detecting ICH in the operating room. The system employs an isocentric C-arm with a flat-panel detector in dual gain mode, correction of x-ray scatter and beam-hardening, and a penalized likelihood (PL) iterative reconstruction method. Performance in ICH detection was investigated using a quantitative phantom focusing on (non-contrast-enhanced) blood-brain contrast, an anthropomorphic head phantom, and a porcine model with injection of fresh blood bolus. The visibility of ICH was characterized in terms of contrast-to-noise ratio (CNR) and qualitative evaluation of images by a neurosurgeon. Across a range of size and contrast of the ICH as well as radiation dose from the CBCT scan, the CNR was found to increase from ~2.2-3.7 for conventional filtered backprojection (FBP) to ~3.9-5.4 for PL at equivalent spatial resolution. The porcine model demonstrated superior ICH detectability for PL. The results support the role of high-quality mobile C-arm CBCT employing advanced reconstruction algorithms for detecting subtle complications in the operating room at lower radiation dose and lower cost than intraoperative CT scanners and/or fixedroom C-arms. Such capability could present a potentially valuable aid to patient safety and QA.

  2. A line fiducial method for geometric calibration of cone-beam CT systems with diverse scan trajectories

    Science.gov (United States)

    Jacobson, M. W.; Ketcha, M. D.; Capostagno, S.; Martin, A.; Uneri, A.; Goerres, J.; De Silva, T.; Reaungamornrat, S.; Han, R.; Manbachi, A.; Stayman, J. W.; Vogt, S.; Kleinszig, G.; Siewerdsen, J. H.

    2018-01-01

    Modern cone-beam CT systems, especially C-arms, are capable of diverse source-detector orbits. However, geometric calibration of these systems using conventional configurations of spherical fiducials (BBs) may be challenged for novel source-detector orbits and system geometries. In part, this is because the BB configurations are designed with careful forethought regarding the intended orbit so that BB marker projections do not overlap in projection views. Examples include helical arrangements of BBs (Rougee et al 1993 Proc. SPIE 1897 161-9) such that markers do not overlap in projections acquired from a circular orbit and circular arrangements of BBs (Cho et al 2005 Med. Phys. 32 968-83). As a more general alternative, this work proposes a calibration method based on an array of line-shaped, radio-opaque wire segments. With this method, geometric parameter estimation is accomplished by relating the 3D line equations representing the wires to the 2D line equations of their projections. The use of line fiducials simplifies many challenges with fiducial recognition and extraction in an orbit-independent manner. For example, their projections can overlap only mildly, for any gantry pose, as long as the wires are mutually non-coplanar in 3D. The method was tested in application to circular and non-circular trajectories in simulation and in real orbits executed using a mobile C-arm prototype for cone-beam CT. Results indicated high calibration accuracy, as measured by forward and backprojection/triangulation error metrics. Triangulation errors on the order of microns and backprojected ray deviations uniformly less than 0.2 mm were observed in both real and simulated orbits. Mean forward projection errors less than 0.1 mm were observed in a comprehensive sweep of different C-arm gantry angulations. Finally, successful integration of the method into a CT imaging chain was demonstrated in head phantom scans.

  3. An autopsy case of otogenic intracranial abscess and meningitis with Bezold's abscess: evaluation of inflammatory bone destruction by postmortem cone-beam CT.

    Science.gov (United States)

    Kanawaku, Yoshimasa; Yanase, Takeshi; Hayashi, Kino; Harada, Kazuki; Kanetake, Jun; Fukunaga, Tatsushige

    2013-11-01

    The deceased was an unidentified young male found unconscious on a walkway. On autopsy, outer and inner fistulae of the left temporal bone, subcutaneous abscess in the left side of the neck and head, and an intracranial abscess were noted. A portion of the left temporal bone was removed and scanned by cone-beam computed tomography (CT) (normally used for dentistry applications) to evaluate the lesion. The cone-beam CT image revealed roughening of the bone wall and hypolucency of the mastoid air cells, consistent with an inflammatory bone lesion. According to autopsy and imaging findings, the cause of death was diagnosed as intracranial abscess with Bezold's abscess secondary to left mastoiditis as a complication of otitis media. Although determining the histopathology of bone specimens is time-consuming and costly work, we believe that use of cone-beam CT for hard tissue specimens can be useful in forensic practice. Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

  4. Investigation of the radiation dose from cone-beam CT for image-guided radiotherapy: A comparison of methodologies.

    Science.gov (United States)

    Buckley, Jarryd G; Wilkinson, Dean; Malaroda, Alessandra; Metcalfe, Peter

    2018-01-01

    Four methodologies were evaluated for quantifying kilovoltage cone-beam computed tomography (CBCT) dose: the Cone-Beam Dose Index (CBDI), IAEA Report 5 recommended methodology (IAEA), the AAPM Task Group 111 methodology (TG111), and the current dose metric; the Computed Tomography Dose Index (CTDI) on two commercial Varian cone-beam CT imaging systems; the Clinac iX On-Board Imager (OBI); and the TrueBeam X-ray Imaging system (XI). The TG111 methodology measured the highest overall dose (21.199 ± 0.035 mGy OBI and 22.420 ± 0.002 XI for pelvis imaging) due to the full scatter of the TG111 phantom and was within 5% of CTDI measurements taken using a full scatter TG111 phantom and 30-cm film strips. CBDI measured the second highest overall dose, within 10% of the TG111, with IAEA measuring the third highest dose. For head CBCT protocols, CBDI measured the highest dose, followed by IAEA. The CTDI method measured lowest across all scan modes highlighting its limitations for CBCT dosimetry. The XI imaging system delivered lower doses for head and thorax scan modes and similar doses to the OBI system for pelvis scan modes due to additional beam hardening filtration in the XI system. The TG111 method measured the highest dose in the center of a CBCT scan during image guidance procedures; however, CBDI provided a good approximation to TG111 with existing CTDI equipment and may be more applicable clinically. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  5. Cone beam CT with zonal filters for simultaneous dose reduction, improved target contrast and automated set-up in radiotherapy

    International Nuclear Information System (INIS)

    Moore, C J; Marchant, T E; Amer, A M

    2006-01-01

    Cone beam CT (CBCT) using a zonal filter is introduced. The aims are reduced concomitant imaging dose to the patient, simultaneous control of body scatter for improved image quality in the tumour target zone and preserved set-up detail for radiotherapy. Aluminium transmission diaphragms added to the CBCT x-ray tube of the Elekta Synergy TM linear accelerator produced an unattenuated beam for a central 'target zone' and a partially attenuated beam for an outer 'set-up zone'. Imaging doses and contrast noise ratios (CNR) were measured in a test phantom for transmission diaphragms 12 and 24 mm thick, for 5 and 10 cm long target zones. The effect on automatic registration of zonal CBCT to conventional CT was assessed relative to full-field and lead-collimated images of an anthropomorphic phantom. Doses along the axis of rotation were reduced by up to 50% in both target and set-up zones, and weighted dose (two thirds surface dose plus one third central dose) was reduced by 10-20% for a 10 cm long target zone. CNR increased by up to 15% in zonally filtered CBCT images compared to full-field images. Automatic image registration remained as robust as that with full-field images and was superior to CBCT coned down using lead-collimation. Zonal CBCT significantly reduces imaging dose and is expected to benefit radiotherapy through improved target contrast, required to assess target coverage, and wide-field edge detail, needed for robust automatic measurement of patient set-up error

  6. Low-contrast visualization in megavoltage cone-beam CT at one beam pulse per projection using thick segmented scintillators

    Science.gov (United States)

    El-Mohri, Youcef; Antonuk, Larry E.; Zhao, Qihua; Choroszucha, Richard B.; Wang, Yi

    2010-04-01

    Megavoltage cone-beam computed tomography (MV CBCT) using an electronic portal imaging device (EPID) is a highly promising technique for providing valuable volumetric information for image guidance in radiotherapy. However, active matrix flat-panel imagers (AMFPIs), which are the established gold standard in portal imaging, require a relatively large dose to create images that are clinically useful. This is a consequence of the inefficiency of the phosphor screens employed in conventional MV AMFPIs, which utilize only ~2% of the incident radiation at 6 MV. Fortunately, the incorporation of thick, segmented scintillators can significantly improve the performance of MV AMFPIs, leading to improved image quality for projection imaging at extremely low dose. It is therefore of interest to explore the performance of such thick scintillators for MV CBCT toward the goal of soft-tissue contrast visualization. In this study, prototype AMFPIs incorporating segmented scintillators based on CsI:Tl and BGO crystals with thicknesses ranging from ~11 to 25 mm have been constructed and evaluated. Each prototype incorporates a detector consisting of a matrix of 120 × 60 scintillator elements separated by reflective septal walls, with an element-to-element pitch of 1.016 mm, coupled to an overlying ~1 mm thick Cu plate. The prototype AMFPIs were incorporated into a bench-top CBCT system, allowing the acquisition of tomographic images of a contrast phantom using a 6 MV radiotherapy photon beam. The phantom consists of a water-equivalent (solid water) cylinder, embedded with tissue-equivalent inserts having electron densities, relative to water, varying from ~0.43 to ~1.47. Reconstructed images of the phantom were obtained down to the lowest available dose (one beam pulse per projection), corresponding to a total scan dose of ~4 cGy using 180 projections. In this article, reconstructed images, contrast, noise and contrast-to-noise ratio for the tissue-equivalent objects using the

  7. A Cone Beam CT-Based Study for Clinical Target Definition Using Pelvic Anatomy During Postprostatectomy Radiotherapy

    International Nuclear Information System (INIS)

    Showalter, Timothy N.; Nawaz, A. Omer; Xiao Ying; Galvin, James M.; Valicenti, Richard K.

    2008-01-01

    Purpose: There are no accepted guidelines for target volume definition for online image-guided radiation therapy (IGRT) after radical prostatectomy (RP). This study used cone beam CT (CBCT) imaging to generate information for use in post-RP IGRT. Methods and Materials: The pelvic anatomy of 10 prostate cancer patients undergoing post-RP radiation therapy (RT) to 68.4 Gy was studied using CBCT images obtained immediately before treatment. Contoured bladder and rectal volumes on CBCT images were compared with planning CT (CT ref ) volumes from seminal vesicle stump (SVS) to bladder-urethral junction. This region was chosen to approximate the prostatic fossa (PF) during a course of post-RP RT. Anterior and posterior planning target volume margins were calculated using ICRU report 71 guidelines, accounting for systematic and random error based on bladder and rectal motion, respectively. Results: A total of 176 CBCT study sets obtained 2 to 5 times weekly were analyzed. The rectal and bladder borders were reliably identified in 166 of 176 (94%) of CBCT images. Relative to CT ref , mean posterior bladder wall position was anterior by 0.1 to 1.5 mm, and mean anterior rectum wall position was posterior by 1.6 to 2.7 mm. Calculated anterior margin as derived from bladder motion ranged from 5.9 to 7.1 mm. Calculated posterior margin as derived from rectal motion ranged from 8.6 to 10.2 mm. Conclusions: Normal tissue anatomy was definable by CBCT imaging throughout the course of post-RP RT, and the interfraction anteroposterior motion of the bladder and rectum was studied. This information should be considered in devising post-RP RT techniques using image guidance

  8. Evaluation of stability of stereotactic space defined by cone-beam CT for the Leksell Gamma Knife Icon.

    Science.gov (United States)

    AlDahlawi, Ismail; Prasad, Dheerendra; Podgorsak, Matthew B

    2017-05-01

    The Gamma Knife Icon comes with an integrated cone-beam CT (CBCT) for image-guided stereotactic treatment deliveries. The CBCT can be used for defining the Leksell stereotactic space using imaging without the need for the traditional invasive frame system, and this allows also for frameless thermoplastic mask stereotactic treatments (single or fractionated) with the Gamma Knife unit. In this study, we used an in-house built marker tool to evaluate the stability of the CBCT-based stereotactic space and its agreement with the standard frame-based stereotactic space. We imaged the tool with a CT indicator box using our CT-simulator at the beginning, middle, and end of the study period (6 weeks) for determining the frame-based stereotactic space. The tool was also scanned with the Icon's CBCT on a daily basis throughout the study period, and the CBCT images were used for determining the CBCT-based stereotactic space. The coordinates of each marker were determined in each CT and CBCT scan using the Leksell GammaPlan treatment planning software. The magnitudes of vector difference between the means of each marker in frame-based and CBCT-based stereotactic space ranged from 0.21 to 0.33 mm, indicating good agreement of CBCT-based and frame-based stereotactic space definition. Scanning 4-month later showed good prolonged stability of the CBCT-based stereotactic space definition. © 2017 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

  9. Technical note: RabbitCT--an open platform for benchmarking 3D cone-beam reconstruction algorithms.

    Science.gov (United States)

    Rohkohl, C; Keck, B; Hofmann, H G; Hornegger, J

    2009-09-01

    Fast 3D cone beam reconstruction is mandatory for many clinical workflows. For that reason, researchers and industry work hard on hardware-optimized 3D reconstruction. Backprojection is a major component of many reconstruction algorithms that require a projection of each voxel onto the projection data, including data interpolation, before updating the voxel value. This step is the bottleneck of most reconstruction algorithms and the focus of optimization in recent publications. A crucial limitation, however, of these publications is that the presented results are not comparable to each other. This is mainly due to variations in data acquisitions, preprocessing, and chosen geometries and the lack of a common publicly available test dataset. The authors provide such a standardized dataset that allows for substantial comparison of hardware accelerated backprojection methods. They developed an open platform RabbitCT (www.rabbitCT.com) for worldwide comparison in backprojection performance and ranking on different architectures using a specific high resolution C-arm CT dataset of a rabbit. This includes a sophisticated benchmark interface, a prototype implementation in C++, and image quality measures. At the time of writing, six backprojection implementations are already listed on the website. Optimizations include multithreading using Intel threading building blocks and OpenMP, vectorization using SSE, and computation on the GPU using CUDA 2.0. There is a need for objectively comparing backprojection implementations for reconstruction algorithms. RabbitCT aims to provide a solution to this problem by offering an open platform with fair chances for all participants. The authors are looking forward to a growing community and await feedback regarding future evaluations of novel software- and hardware-based acceleration schemes.

  10. Automated double-cone-beam CT fusion technique. Enhanced evaluation of glue distribution in cases of spinal dural arteriovenous fistula (SDAVF) embolisation

    Energy Technology Data Exchange (ETDEWEB)

    Farago, Giuseppe [Foundation Neurological Institute ' ' C. Besta' ' , Department of Interventional Neuroradiology, Milan (Italy); Fondazione IRCCS Istituto Neurologico Carlo Besta, Department of Interventional Neuroradiology, Milan (Italy); Caldiera, V. [Foundation Neurological Institute ' ' C. Besta' ' , Department of Interventional Neuroradiology, Milan (Italy); Antozzi, C.; Bellino, A. [Foundation Neurological Institute ' ' C. Besta' ' , Department of Neuroimmunology and Neuromuscular Diseases, Milan (Italy); Innocenti, A. [Foundation Neurological Institute ' ' C. Besta' ' , Department of Neuro-Oncology, Milan (Italy); Ciceri, E. [Foundation Neurological Institute ' ' C. Besta' ' , Department of Interventional Neuroradiology, Milan (Italy); Azienda Ospedaliera Universitaria Integrata Borgo Trento, Department of Neuroradiology, Verona (Italy)

    2017-05-15

    Spinal dural arteriovenous fistulas (SDAVFs) are acquired diseases that represent the majority of all arteriovenous spinal shunts, leading to progressive and disabling myelopathy. Treatment is focused on accurately disconnecting the fistula point. We present our experience with the double-cone-beam CT fusion technique successfully applied to evaluate treatment results in a series of SDAVFs. Between November 2011 and December 2015 we performed double-DynaCT acquisition (pre- and post-embolisation) in 12 cases of SDAVF. A successful DynaCT fusion technique was only achieved in the group of patients with pre- and post-treatment images acquired at the same time as the treatment session, under general anaesthesia (4/12). DynaCT performed on different days proved to be inadequate for the automated fusion technique because of changes in the body position (8/12). A pre-treatment flat-panel cone-beam CT with contrast, at the time of diagnostic angiography, can be very helpful to detect the correct level of the fistula and the relationship between the fistula and the surrounding structures. In case of the endovascular approach, additional post-treatment native acquisition merged with the pre-treatment acquisition (double-cone-beam CT fusion technique) permits to immediately evaluate the distribution of the glue cast and to confirm the success of the procedure. (orig.)

  11. Mechanosensitive gating of Kv channels.

    Directory of Open Access Journals (Sweden)

    Catherine E Morris

    Full Text Available K-selective voltage-gated channels (Kv are multi-conformation bilayer-embedded proteins whose mechanosensitive (MS Popen(V implies that at least one conformational transition requires the restructuring of the channel-bilayer interface. Unlike Morris and colleagues, who attributed MS-Kv responses to a cooperative V-dependent closed-closed expansion↔compaction transition near the open state, Mackinnon and colleagues invoke expansion during a V-independent closed↔open transition. With increasing membrane tension, they suggest, the closed↔open equilibrium constant, L, can increase >100-fold, thereby taking steady-state Popen from 0→1; "exquisite sensitivity to small…mechanical perturbations", they state, makes a Kv "as much a mechanosensitive…as…a voltage-dependent channel". Devised to explain successive gK(V curves in excised patches where tension spontaneously increased until lysis, their L-based model falters in part because of an overlooked IK feature; with recovery from slow inactivation factored in, their g(V datasets are fully explained by the earlier model (a MS V-dependent closed-closed transition, invariant L≥4. An L-based MS-Kv predicts neither known Kv time courses nor the distinctive MS responses of Kv-ILT. It predicts Kv densities (hence gating charge per V-sensor several-fold different from established values. If opening depended on elevated tension (L-based model, standard gK(V operation would be compromised by animal cells' membrane flaccidity. A MS V-dependent transition is, by contrast, unproblematic on all counts. Since these issues bear directly on recent findings that mechanically-modulated Kv channels subtly tune pain-related excitability in peripheral mechanoreceptor neurons we undertook excitability modeling (evoked action potentials. Kvs with MS V-dependent closed-closed transitions produce nuanced mechanically-modulated excitability whereas an L-based MS-Kv yields extreme, possibly excessive

  12. SU-F-T-215: An Investigation Of Multi-Scanner CT Hounsfield Unit Calibration for Pencil Beam Scanning Proton Therapy Using 3D Gamma Analysis

    International Nuclear Information System (INIS)

    Zhang, J; Li, X; Liu, G; Liu, Q; Liang, J; Ding, X

    2016-01-01

    Purpose: We compare and investigate the dosimetric impacts on pencil beam scanning (PBS) proton treatment plans generated with CT calibration curves from four different CT scanners and one averaged ‘global’ CT calibration curve. Methods: The four CT scanners are located at three different hospital locations within the same health system. CT density calibration curves were collected from these scanners using the same CT calibration phantom and acquisition parameters. Mass density to HU value tables were then commissioned in a commercial treatment planning system. Five disease sites were chosen for dosimetric comparisons at brain, lung, head and neck, adrenal, and prostate. Three types of PBS plans were generated at each treatment site using SFUD, IMPT, and robustness optimized IMPT techniques. 3D dose differences were investigated using 3D Gamma analysis. Results: The CT calibration curves for all four scanners display very similar shapes. Large HU differences were observed at both the high HU and low HU regions of the curves. Large dose differences were generally observed at the distal edges of the beams and they are beam angle dependent. Out of the five treatment sites, lung plans exhibits the most overall range uncertainties and prostate plans have the greatest dose discrepancy. There are no significant differences between the SFUD, IMPT, and the RO-IMPT methods. 3D gamma analysis with 3%, 3 mm criteria showed all plans with greater than 95% passing rate. Two of the scanners with close HU values have negligible dose difference except for lung. Conclusion: Our study shows that there are more than 5% dosimetric differences between different CT calibration curves. PBS treatment plans generated with SFUD, IMPT, and the robustness optimized IMPT has similar sensitivity to the CT density uncertainty. More patient data and tighter gamma criteria based on structure location and size will be used for further investigation.

  13. Quantitative assessment of the physical potential of proton beam range verification with PET/CT

    NARCIS (Netherlands)

    Knopf, A; Parodi, K.; Paganetti, Harald; Lo Cascio, E; Bonab, A; Bortfeld, Thomas

    2008-01-01

    A recent clinical pilot study demonstrated the feasibility of offline PET/CT range verification for proton therapy treatments. In vivo PET measurements are challenged by blood perfusion, variations of tissue compositions, patient motion and image co-registration uncertainties. Besides these

  14. Implant planning and placement using optical scanning and cone beam CT technology

    NARCIS (Netherlands)

    van der Zel, J.M.

    2008-01-01

    There is a growing interest in minimally invasive implant therapy as a standard prosthodontic treatment, providing complete restoration of occlusal function. A new treatment method (CADDIMA), which combines both computerized tomographic (CT) and optical laser-scan data for planning and design of

  15. A graphical approach to optimizing variable-kernel smoothing parameters for improved deformable registration of CT and cone beam CT images

    Science.gov (United States)

    Hart, Vern; Burrow, Damon; Li, X. Allen

    2017-08-01

    A systematic method is presented for determining optimal parameters in variable-kernel deformable image registration of cone beam CT and CT images, in order to improve accuracy and convergence for potential use in online adaptive radiotherapy. Assessed conditions included the noise constant (symmetric force demons), the kernel reduction rate, the kernel reduction percentage, and the kernel adjustment criteria. Four such parameters were tested in conjunction with reductions of 5, 10, 15, 20, 30, and 40%. Noise constants ranged from 1.0 to 1.9 for pelvic images in ten prostate cancer patients. A total of 516 tests were performed and assessed using the structural similarity index. Registration accuracy was plotted as a function of iteration number and a least-squares regression line was calculated, which implied an average improvement of 0.0236% per iteration. This baseline was used to determine if a given set of parameters under- or over-performed. The most accurate parameters within this range were applied to contoured images. The mean Dice similarity coefficient was calculated for bladder, prostate, and rectum with mean values of 98.26%, 97.58%, and 96.73%, respectively; corresponding to improvements of 2.3%, 9.8%, and 1.2% over previously reported values for the same organ contours. This graphical approach to registration analysis could aid in determining optimal parameters for Demons-based algorithms. It also establishes expectation values for convergence rates and could serve as an indicator of non-physical warping, which often occurred in cases  >0.6% from the regression line.

  16. Design of 400 kV intense neutron generator

    International Nuclear Information System (INIS)

    Lu Xiaolong; Yang Yao; Yan Siqi; He Xiongying; Yan Xiaojun; Yao Ze'en

    2012-01-01

    The design of 400 kV intense neutron generator was presented. The electric fields of high voltage electrode and accelerating column were simulated using Poisson/Superfish code. The results show that the maximum value of electric field in the key areas is far less than the value of breakdown electric field. The program for simulating high current ion beam envelope, called IONB1.0, was developed based on high current ion beam envelope equations. The envelopes of 40 mA D ion beam were simulated in the beam transport system of 400 kV intense neutron generator. The results show that the strong focus performance from the accelerating column with two gaps high gradient electric field can effectively offset ion beam divergence caused by the space charge effect. The radius of envelope is about 3 cm at accelerating column exit. The beam transport system composed of a space charge lens and a triple quadrupole magnetic lens can focused 40 mA D beam on the target at 140 cm distance, and the beam spot diameter is less than 2 cm. (authors)

  17. High-quality 3D correction of ring and radiant artifacts in flat panel detector-based cone beam volume CT imaging

    Science.gov (United States)

    Abu Anas, Emran Mohammad; Kim, Jae Gon; Lee, Soo Yeol; Kamrul Hasan, Md

    2011-10-01

    The use of an x-ray flat panel detector is increasingly becoming popular in 3D cone beam volume CT machines. Due to the deficient semiconductor array manufacturing process, the cone beam projection data are often corrupted by different types of abnormalities, which cause severe ring and radiant artifacts in a cone beam reconstruction image, and as a result, the diagnostic image quality is degraded. In this paper, a novel technique is presented for the correction of error in the 2D cone beam projections due to abnormalities often observed in 2D x-ray flat panel detectors. Template images are derived from the responses of the detector pixels using their statistical properties and then an effective non-causal derivative-based detection algorithm in 2D space is presented for the detection of defective and mis-calibrated detector elements separately. An image inpainting-based 3D correction scheme is proposed for the estimation of responses of defective detector elements, and the responses of the mis-calibrated detector elements are corrected using the normalization technique. For real-time implementation, a simplification of the proposed off-line method is also suggested. Finally, the proposed algorithms are tested using different real cone beam volume CT images and the experimental results demonstrate that the proposed methods can effectively remove ring and radiant artifacts from cone beam volume CT images compared to other reported techniques in the literature.

  18. High-quality 3D correction of ring and radiant artifacts in flat panel detector-based cone beam volume CT imaging

    International Nuclear Information System (INIS)

    Anas, Emran Mohammad Abu; Hasan, Md Kamrul; Kim, Jae Gon; Lee, Soo Yeol

    2011-01-01

    The use of an x-ray flat panel detector is increasingly becoming popular in 3D cone beam volume CT machines. Due to the deficient semiconductor array manufacturing process, the cone beam projection data are often corrupted by different types of abnormalities, which cause severe ring and radiant artifacts in a cone beam reconstruction image, and as a result, the diagnostic image quality is degraded. In this paper, a novel technique is presented for the correction of error in the 2D cone beam projections due to abnormalities often observed in 2D x-ray flat panel detectors. Template images are derived from the responses of the detector pixels using their statistical properties and then an effective non-causal derivative-based detection algorithm in 2D space is presented for the detection of defective and mis-calibrated detector elements separately. An image inpainting-based 3D correction scheme is proposed for the estimation of responses of defective detector elements, and the responses of the mis-calibrated detector elements are corrected using the normalization technique. For real-time implementation, a simplification of the proposed off-line method is also suggested. Finally, the proposed algorithms are tested using different real cone beam volume CT images and the experimental results demonstrate that the proposed methods can effectively remove ring and radiant artifacts from cone beam volume CT images compared to other reported techniques in the literature.

  19. Recent developments with a prototype fan-beam optical CT scanner

    Science.gov (United States)

    Campbell, W. G.; Jirasek, A.; Wells, D.

    2013-06-01

    The latest design of a prototype fan-beam optical computed tomography scanner is presented. A new beam creation system consists of a 635 nm laser diode module with variable, DC voltage-controlled beam intensity. A change in scanner alignment allows for the elimination of ring artefacts caused by data corruption that is spaced symmetrically across the detector array. These artefacts, as well as a pair of streaking artefacts caused by flask seams, are removed in sinogram space. A flask registration technique has been developed that allows for accurate, reproducible dosimeter placement. Protocol investigations with gel dosimeters have indicated the importance of: i) proper cooling techniques during gel manufacture, and ii) scanning the dosimeter while it is at room temperature. Latest reconstructions of a normoxic polymer gel dosimeter are presented as an indicator of current system performance.

  20. Recent developments with a prototype fan-beam optical CT scanner

    International Nuclear Information System (INIS)

    Campbell, W G; Jirasek, A; Wells, D

    2013-01-01

    The latest design of a prototype fan-beam optical computed tomography scanner is presented. A new beam creation system consists of a 635 nm laser diode module with variable, DC voltage-controlled beam intensity. A change in scanner alignment allows for the elimination of ring artefacts caused by data corruption that is spaced symmetrically across the detector array. These artefacts, as well as a pair of streaking artefacts caused by flask seams, are removed in sinogram space. A flask registration technique has been developed that allows for accurate, reproducible dosimeter placement. Protocol investigations with gel dosimeters have indicated the importance of: i) proper cooling techniques during gel manufacture, and ii) scanning the dosimeter while it is at room temperature. Latest reconstructions of a normoxic polymer gel dosimeter are presented as an indicator of current system performance.

  1. Simulation-aided investigation of beam hardening induced errors in CT dimensional metrology

    DEFF Research Database (Denmark)

    Tan, Ye; Kiekens, Kim; Welkenhuyzen, Frank

    2014-01-01

    are mutually correlated, it remains challenging to interpret measurement results and to identify the distinct error sources. Since simulations allow isolating the different affecting factors, they form a useful complement to experimental investigations. Dewulf et al (2012 CIRP Ann. Manuf. Technol. 61 495......–8) investigated the influence of beam hardening correction parameters on the diameter of a calibrated steel pin in different experimental set-ups. It was clearly shown that an inappropriate beam hardening correction can result in significant dimensional errors. This paper confirms these results using simulations...... of a pin surrounded by a stepped cylinder: a clear discontinuity in the measured diameter of the inner pin is observed where it enters the surrounding material. The results are expanded with an investigation of the beam hardening effect on the measurement results for both inner and outer diameters...

  2. Simulation-aided investigation of beam hardening induced errors in CT dimensional metrology

    DEFF Research Database (Denmark)

    Tan, Ye; Kiekens, Kim; Welkenhuyzen, Frank

    2013-01-01

    most of these factors are mutually correlated, it remains challenging to interpret measurement results and to identify the distinct error sources. Since simulations allow isolating the different affecting factors, they form a useful complement to experimental investigations. Dewulf et.al [5......] investigated the influence of beam hardening correction parameters on the diameter of a calibrated steel pin in different experimental set-ups. It was clearly shown that inappropriate beam hardening correction can result in significant dimensional errors. This paper confirms these results using simulations...... of a pin surrounded by a stepped cylinder: a clear discontinuity in the measured diameter of the inner pin is observed where it enters the surrounding material. The results are expanded with an investigation of the beam hardening effect on the measurement results for both inner and outer diameters...

  3. A BPF-FBP tandem algorithm for image reconstruction in reverse helical cone-beam CT

    International Nuclear Information System (INIS)

    Cho, Seungryong; Xia, Dan; Pellizzari, Charles A.; Pan Xiaochuan

    2010-01-01

    Purpose: Reverse helical cone-beam computed tomography (CBCT) is a scanning configuration for potential applications in image-guided radiation therapy in which an accurate anatomic image of the patient is needed for image-guidance procedures. The authors previously developed an algorithm for image reconstruction from nontruncated data of an object that is completely within the reverse helix. The purpose of this work is to develop an image reconstruction approach for reverse helical CBCT of a long object that extends out of the reverse helix and therefore constitutes data truncation. Methods: The proposed approach comprises of two reconstruction steps. In the first step, a chord-based backprojection-filtration (BPF) algorithm reconstructs a volumetric image of an object from the original cone-beam data. Because there exists a chordless region in the middle of the reverse helix, the image obtained in the first step contains an unreconstructed central-gap region. In the second step, the gap region is reconstructed by use of a Pack-Noo-formula-based filteredbackprojection (FBP) algorithm from the modified cone-beam data obtained by subtracting from the original cone-beam data the reprojection of the image reconstructed in the first step. Results: The authors have performed numerical studies to validate the proposed approach in image reconstruction from reverse helical cone-beam data. The results confirm that the proposed approach can reconstruct accurate images of a long object without suffering from data-truncation artifacts or cone-angle artifacts. Conclusions: They developed and validated a BPF-FBP tandem algorithm to reconstruct images of a long object from reverse helical cone-beam data. The chord-based BPF algorithm was utilized for converting the long-object problem into a short-object problem. The proposed approach is applicable to other scanning configurations such as reduced circular sinusoidal trajectories.

  4. SU-F-J-32: Do We Need KV Imaging During CBCT Based Patient Set-Up for Lung Radiation Therapy?

    Energy Technology Data Exchange (ETDEWEB)

    Gopal, A; Zhou, J; Prado, K; D’souza, W; Yi, B [University of Maryland School of Medicine, Baltimore, MD (United States)

    2016-06-15

    Purpose: To evaluate the role of 2D kilovoltage (kV) imaging to complement cone beam CT (CBCT) imaging in a shift threshold based image guided radiation therapy (IGRT) strategy for conventional lung radiotherapy. Methods: A retrospective study was conducted by analyzing IGRT couch shift trends for 15 patients that received lung radiation therapy to evaluate the benefit of performing orthogonal kV imaging prior to CBCT imaging. Herein, a shift threshold based IGRT protocol was applied, which would mandate additional CBCT verification if the applied patient shifts exceeded 3 mm to avoid intraobserver variability in CBCT registration and to confirm table shifts. For each patient, two IGRT strategies: kV + CBCT and CBCT alone, were compared and the recorded patient shifts were categorized into whether additional CBCT acquisition would have been mandated or not. The effectiveness of either strategy was gauged by the likelihood of needing additional CBCT imaging for accurate patient set-up. Results: The use of CBCT alone was 6 times more likely to require an additional CBCT than KV+CBCT, for a 3 mm shift threshold (88% vs 14%). The likelihood of additional CBCT verification generally increased with lower shift thresholds, and was significantly lower when kV+CBCT was used (7% with 5 mm shift threshold, 36% with 2 mm threshold), than with CBCT alone (61% with 5 mm shift threshold, 97% with 2 mm threshold). With CBCT alone, treatment time increased by 2.2 min and dose increased by 1.9 cGy per fraction on average due to additional CBCT with a 3mm shift threshold. Conclusion: The benefit of kV imaging to screen for gross misalignments led to more accurate CBCT based patient localization compared with using CBCT alone. The subsequently reduced need for additional CBCT verification will minimize treatment time and result in less overall patient imaging dose.

  5. Ultrafast and scalable cone-beam CT reconstruction using MapReduce in a cloud computing environment.

    Science.gov (United States)

    Meng, Bowen; Pratx, Guillem; Xing, Lei

    2011-12-01

    Four-dimensional CT (4DCT) and cone beam CT (CBCT) are widely used in radiation therapy for accurate tumor target definition and localization. However, high-resolution and dynamic image reconstruction is computationally demanding because of the large amount of data processed. Efficient use of these imaging techniques in the clinic requires high-performance computing. The purpose of this work is to develop a novel ultrafast, scalable and reliable image reconstruction technique for 4D CBCT∕CT using a parallel computing framework called MapReduce. We show the utility of MapReduce for solving large-scale medical physics problems in a cloud computing environment. In this work, we accelerated the Feldcamp-Davis-Kress (FDK) algorithm by porting it to Hadoop, an open-source MapReduce implementation. Gated phases from a 4DCT scans were reconstructed independently. Following the MapReduce formalism, Map functions were used to filter and backproject subsets of projections, and Reduce function to aggregate those partial backprojection into the whole volume. MapReduce automatically parallelized the reconstruction process on a large cluster of computer nodes. As a validation, reconstruction of a digital phantom and an acquired CatPhan 600 phantom was performed on a commercial cloud computing environment using the proposed 4D CBCT∕CT reconstruction algorithm. Speedup of reconstruction time is found to be roughly linear with the number of nodes employed. For instance, greater than 10 times speedup was achieved using 200 nodes for all cases, compared to the same code executed on a single machine. Without modifying the code, faster reconstruction is readily achievable by allocating more nodes in the cloud computing environment. Root mean square error between the images obtained using MapReduce and a single-threaded reference implementation was on the order of 10(-7). Our study also proved that cloud computing with MapReduce is fault tolerant: the reconstruction completed

  6. TH-A-18C-04: Ultrafast Cone-Beam CT Scatter Correction with GPU-Based Monte Carlo Simulation

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Y [UT Southwestern Medical Center, Dallas, TX (United States); Southern Medical University, Guangzhou (China); Bai, T [UT Southwestern Medical Center, Dallas, TX (United States); Xi' an Jiaotong University, Xi' an (China); Yan, H; Ouyang, L; Wang, J; Pompos, A; Jiang, S; Jia, X [UT Southwestern Medical Center, Dallas, TX (United States); Zhou, L [Southern Medical University, Guangzhou (China)

    2014-06-15

    Purpose: Scatter artifacts severely degrade image quality of cone-beam CT (CBCT). We present an ultrafast scatter correction framework by using GPU-based Monte Carlo (MC) simulation and prior patient CT image, aiming at automatically finish the whole process including both scatter correction and reconstructions within 30 seconds. Methods: The method consists of six steps: 1) FDK reconstruction using raw projection data; 2) Rigid Registration of planning CT to the FDK results; 3) MC scatter calculation at sparse view angles using the planning CT; 4) Interpolation of the calculated scatter signals to other angles; 5) Removal of scatter from the raw projections; 6) FDK reconstruction using the scatter-corrected projections. In addition to using GPU to accelerate MC photon simulations, we also use a small number of photons and a down-sampled CT image in simulation to further reduce computation time. A novel denoising algorithm is used to eliminate MC scatter noise caused by low photon numbers. The method is validated on head-and-neck cases with simulated and clinical data. Results: We have studied impacts of photo histories, volume down sampling factors on the accuracy of scatter estimation. The Fourier analysis was conducted to show that scatter images calculated at 31 angles are sufficient to restore those at all angles with <0.1% error. For the simulated case with a resolution of 512×512×100, we simulated 10M photons per angle. The total computation time is 23.77 seconds on a Nvidia GTX Titan GPU. The scatter-induced shading/cupping artifacts are substantially reduced, and the average HU error of a region-of-interest is reduced from 75.9 to 19.0 HU. Similar results were found for a real patient case. Conclusion: A practical ultrafast MC-based CBCT scatter correction scheme is developed. The whole process of scatter correction and reconstruction is accomplished within 30 seconds. This study is supported in part by NIH (1R01CA154747-01), The Core Technology Research

  7. Parameter selection in limited data cone-beam CT reconstruction using edge-preserving total variation algorithms

    Science.gov (United States)

    Lohvithee, Manasavee; Biguri, Ander; Soleimani, Manuchehr

    2017-12-01

    There are a number of powerful total variation (TV) regularization methods that have great promise in limited data cone-beam CT reconstruction with an enhancement of image quality. These promising TV methods require careful selection of the image reconstruction parameters, for which there are no well-established criteria. This paper presents a comprehensive evaluation of parameter selection in a number of major TV-based reconstruction algorithms. An appropriate way of selecting the values for each individual parameter has been suggested. Finally, a new adaptive-weighted projection-controlled steepest descent (AwPCSD) algorithm is presented, which implements the edge-preserving function for CBCT reconstruction with limited data. The proposed algorithm shows significant robustness compared to three other existing algorithms: ASD-POCS, AwASD-POCS and PCSD. The proposed AwPCSD algorithm is able to preserve the edges of the reconstructed images better with fewer sensitive parameters to tune.

  8. Comparison of a dental cone beam CT with a multi-detector row CT on effective doses and physical image quality

    International Nuclear Information System (INIS)

    Yoshida, Yutaka; Tokumori, Kenji; Okamura, Kazutoshi; Yoshiura, Kazunori

    2011-01-01

    The purpose of this study was to compare a dental cone beam computed tomography (dental CBCT) and a multi-detector row CT (MDCT) using effective doses and physical image quality. A dental mode (D-mode) and an implant mode (I-mode) were employed for calculating effective doses. Field of view (FOV) size of the MDCT was 150 mm. Three types of images were obtained using 3 different reconstruction functions: FC1 (for abdomen images), FC30 (for internal ear and bone images) and FC81 (for high resolution images). Effective doses obtained with the D-mode and with the I-mode were about 20% and 50% of those obtained with the MDCT, respectively. Resolution properties obtained with the D-mode and I-mode were superior to that of the MDCT in a high frequency range. Noise properties of the D-mode and the I-mode were better than those with FC81. It was found that the dental CBCT has better potential as compared with MDCT in both dental and implant modes. (author)

  9. Time-resolved C-arm cone beam CT angiography (TR-CBCTA) imaging from a single short-scan C-arm cone beam CT acquisition with intra-arterial contrast injection

    Science.gov (United States)

    Li, Yinsheng; Garrett, John W.; Li, Ke; Wu, Yijing; Johnson, Kevin; Schafer, Sebastian; Strother, Charles; Chen, Guang-Hong

    2018-04-01

    Time-resolved C-arm cone-beam CT (CBCT) angiography (TR-CBCTA) images can be generated from a series of CBCT acquisitions that satisfy data sufficiency condition in analytical image reconstruction theory. In this work, a new technique was developed to generate TR-CBCTA images from a single short-scan CBCT data acquisition with contrast media injection. The reconstruction technique enabling this application is a previously developed image reconstruction technique, synchronized multi-artifact reduction with tomographic reconstruction (SMART-RECON). In this new application, the acquired short-scan CBCT projection data were sorted into a union of several sub-sectors of view angles and each sub-sector of view angles corresponds to an individual image volume to be reconstructed. The SMART-RECON method was then used to jointly reconstruct all of these individual image volumes under two constraints: (1) each individual image volume is maximally consistent with the measured cone-beam projection data within the corresponding view angle sector and (2) the nuclear norm of the image matrix is minimized. The difference between these reconstructed individual image volumes is used to generated the desired subtracted angiograms. To validate the technique, numerical simulation data generated from a fractal tree angiogram phantom were used to quantitatively study the accuracy of the proposed method and retrospective in vivo human subject studies were used to demonstrate the feasibility of generating TR-CBCTA in clinical practice.

  10. Het effect van Cone Beam CT op het behandelresultaat in de mondzorg

    NARCIS (Netherlands)

    Stoop P; van Dillen T; BMT

    2012-01-01

    Cone beam Computed Tomography (CBCT) is een relatief nieuwe techniek die met röntgenstraling driedimensionale afbeeldingen van de mond maakt en wordt gebruikt om diagnoses te stellen. In Nederland is deze techniek binnen de mondzorg in opkomst. Uit onderzoek van het RIVM blijkt dat er geen goede

  11. Comparative diagnostic yield of cone beam CT reconstruction using various software programs on the detection of vertical root fractures.

    Science.gov (United States)

    Melo, S L S; Haiter-Neto, F; Correa, L R; Scarfe, W C; Farman, A G

    2013-01-01

    To evaluate the effect on diagnostic yield in the detection of experimentally induced vertical root fractures on cone beam CT images using four dental software program. 190 single-rooted extracted human teeth were divided into three groups according to the pulp canal status: unrestored (UR), filled with gutta-percha (GP) and restored with a metallic custom post (Post). One-half of the sample of each group was artificially fractured and the segments repositioned. All teeth were scanned on a cone beam CT device at 0.2 mm nominal voxel resolution (i-CAT Platinum; Imaging Sciences International, Hatfield, PA). The data were exported as digital imaging and communications in medicine files and imported into Dolphin Imaging & Management Solutions, v. 11.5 (Patterson Dental Supply Inc., St Paul, MN), InVivoDental, v. 5.0 (Anatomage Inc., San Jose, CA) and Kodak Dental Imaging Software 3D module, v. 2.1.11 (Carestream Health Inc., Rochester, NY) software. Cross-sectional images in the acquisition (using Xoran CAT™, v. 3.0.34 software; Xoran Technologies, Ann Arbor, MI) and additional software were presented to three calibrated oral radiologists who rated the presence or absence of root fracture on a five-point scale. Receiver operating characteristic analysis was performed, and treatment comparisons compared by analysis of variance and pairwise comparisons were performed using Tukey's test at an a priori value of α < 0.05%. All dental software performed equally at detecting fractures. Fractures were significantly more difficult to detect when posts were present. The diagnosis of root fracture is software-independent. The presence of an intracanal metallic post significantly decreases the detection of artificially created root fractures.

  12. Morphology of bone defects in patient with unilateral cleft lip and palate. Cone beam x-ray CT evaluation

    International Nuclear Information System (INIS)

    Kyo, Iyu; Kubota, Masato; Sato, Yuki; Nakano, Haruhisa; Maki, Koutaro

    2006-01-01

    Orthodontic treatment planning of the cleft lip and palate vary according to the morphology of the alveolar bone and palatal bone. The purpose of this study is to evaluate the three-dimensional anatomy of the alveolar and palatal bone in children with complete unilateral cleft lip and palate. Thirty-three nonsyndromic consecutive patients with complete unilateral cleft lip and palate were treated by the cleft palate team at Showa University. Each patient had lip and palate surgeries at Showa University. Cone beam CT radiographs (CB MercuRay, Hitachi) were taken prior to secondary bone grafting, and were classified according to the method of Kita et al. 1997. Cone beam CT radiographs showed multiple types of alveolar and palatal bone morphology, and focused on special types described in the method of Kita et al. It was most frequently found that bone defects in the alveolar crest showed similar patterns in both buccal and palatal aspect, and the buccal bone defect in the nasal floor was larger than the palatal bone defect in the nasal floor. In 80% of the patients, the palatal bone defect showed similar patterns in both anterior and posterior aspects, and the anterior palatal bone defect was smaller than the posterior palatal bone defect. In addition, inadequate bone bridges were frequently found at the cleft site. It is suggested that patients with unilateral cleft lip and palate have various types of alveolar and palatal bone morphology, and are required to take three-dimensional radiographic X-rays prior to any orthodontic treatment. (author)

  13. Five-dimensional motion compensation for respiratory and cardiac motion with cone-beam CT of the thorax region

    Science.gov (United States)

    Sauppe, Sebastian; Hahn, Andreas; Brehm, Marcus; Paysan, Pascal; Seghers, Dieter; Kachelrieß, Marc

    2016-03-01

    We propose an adapted method of our previously published five-dimensional (5D) motion compensation (MoCo) algorithm1, developed for micro-CT imaging of small animals, to provide for the first time motion artifact-free 5D cone-beam CT (CBCT) images from a conventional flat detector-based CBCT scan of clinical patients. Image quality of retrospectively respiratory- and cardiac-gated volumes from flat detector CBCT scans is deteriorated by severe sparse projection artifacts. These artifacts further complicate motion estimation, as it is required for MoCo image reconstruction. For high quality 5D CBCT images at the same x-ray dose and the same number of projections as todays 3D CBCT we developed a double MoCo approach based on motion vector fields (MVFs) for respiratory and cardiac motion. In a first step our already published four-dimensional (4D) artifact-specific cyclic motion-compensation (acMoCo) approach is applied to compensate for the respiratory patient motion. With this information a cyclic phase-gated deformable heart registration algorithm is applied to the respiratory motion-compensated 4D CBCT data, thus resulting in cardiac MVFs. We apply these MVFs on double-gated images and thereby respiratory and cardiac motion-compensated 5D CBCT images are obtained. Our 5D MoCo approach processing patient data acquired with the TrueBeam 4D CBCT system (Varian Medical Systems). Our double MoCo approach turned out to be very efficient and removed nearly all streak artifacts due to making use of 100% of the projection data for each reconstructed frame. The 5D MoCo patient data show fine details and no motion blurring, even in regions close to the heart where motion is fastest.

  14. Motion vector field phase-to-amplitude resampling for 4D motion-compensated cone-beam CT

    Science.gov (United States)

    Sauppe, Sebastian; Kuhm, Julian; Brehm, Marcus; Paysan, Pascal; Seghers, Dieter; Kachelrieß, Marc

    2018-02-01

    We propose a phase-to-amplitude resampling (PTAR) method to reduce motion blurring in motion-compensated (MoCo) 4D cone-beam CT (CBCT) image reconstruction, without increasing the computational complexity of the motion vector field (MVF) estimation approach. PTAR is able to improve the image quality in reconstructed 4D volumes, including both regular and irregular respiration patterns. The PTAR approach starts with a robust phase-gating procedure for the initial MVF estimation and then switches to a phase-adapted amplitude gating method. The switch implies an MVF-resampling, which makes them amplitude-specific. PTAR ensures that the MVFs, which have been estimated on phase-gated reconstructions, are still valid for all amplitude-gated reconstructions. To validate the method, we use an artificially deformed clinical CT scan with a realistic breathing pattern and several patient data sets acquired with a TrueBeamTM integrated imaging system (Varian Medical Systems, Palo Alto, CA, USA). Motion blurring, which still occurs around the area of the diaphragm or at small vessels above the diaphragm in artifact-specific cyclic motion compensation (acMoCo) images based on phase-gating, is significantly reduced by PTAR. Also, small lung structures appear sharper in the images. This is demonstrated both for simulated and real patient data. A quantification of the sharpness of the diaphragm confirms these findings. PTAR improves the image quality of 4D MoCo reconstructions compared to conventional phase-gated MoCo images, in particular for irregular breathing patterns. Thus, PTAR increases the robustness of MoCo reconstructions for CBCT. Because PTAR does not require any additional steps for the MVF estimation, it is computationally efficient. Our method is not restricted to CBCT but could rather be applied to other image modalities.

  15. 4D cone beam CT-based dose assessment for SBRT lung cancer treatment

    International Nuclear Information System (INIS)

    Cai, Weixing; Dhou, Salam; Cifter, Fulya; Myronakis, Marios; Hurwitz, Martina H; Williams, Christopher L; Berbeco, Ross I; Seco, Joao; Lewis, John H

    2016-01-01

    The purpose of this research is to develop a 4DCBCT-based dose assessment method for calculating actual delivered dose for patients with significant respiratory motion or anatomical changes during the course of SBRT. To address the limitation of 4DCT-based dose assessment, we propose to calculate the delivered dose using time-varying (‘fluoroscopic’) 3D patient images generated from a 4DCBCT-based motion model. The method includes four steps: (1) before each treatment, 4DCBCT data is acquired with the patient in treatment position, based on which a patient-specific motion model is created using a principal components analysis algorithm. (2) During treatment, 2D time-varying kV projection images are continuously acquired, from which time-varying ‘fluoroscopic’ 3D images of the patient are reconstructed using the motion model. (3) Lateral truncation artifacts are corrected using planning 4DCT images. (4) The 3D dose distribution is computed for each timepoint in the set of 3D fluoroscopic images, from which the total effective 3D delivered dose is calculated by accumulating deformed dose distributions. This approach is validated using six modified XCAT phantoms with lung tumors and different respiratory motions derived from patient data. The estimated doses are compared to that calculated using ground-truth XCAT phantoms. For each XCAT phantom, the calculated delivered tumor dose values generally follow the same trend as that of the ground truth and at most timepoints the difference is less than 5%. For the overall delivered dose, the normalized error of calculated 3D dose distribution is generally less than 3% and the tumor D95 error is less than 1.5%. XCAT phantom studies indicate the potential of the proposed method to accurately estimate 3D tumor dose distributions for SBRT lung treatment based on 4DCBCT imaging and motion modeling. Further research is necessary to investigate its performance for clinical patient data. (paper)

  16. Radiologic study of the healing process of the extracted socket of beagle dogs using cone beam CT

    Energy Technology Data Exchange (ETDEWEB)

    Cho, Bong Won; Lee, Won; Lee, Byung Do [Department of Oral and Maxillofacial Radiology, School of Dentistry, Wonkwang University, Iksan (Korea, Republic of); Kim, De Sok [Department of Electrical Engineering, Korea Advanced Institute of Science and Technology, Daejeon (Korea, Republic of)

    2009-03-15

    To longitudinally observe the healing process of extracted socket and the alterations of the residual ridge in healthy adult dogs using cone beam CT (CBCT). The mandibular premolars of two beagle dogs were removed and the extraction sites were covered with the gingival tissue. CBCTs (3D X-ray CT scanner, Alphard vega, Asahi Co.) were taken at baseline and at 1 week interval for 12 weeks. Radiographic density of extracted wounds was measured on normalized images with a custom-made image analysis program. The amount of alveolar crestal resorption after the teeth extraction was measured with a reformatted three-dimensional image using CBCT. Bony healing pattern of extracted wound of each group was also longitudinally observed and analyzed. Dimensional changes occurred during the first 6 weeks following the extraction of dogs' mandibular premolars. The reduction of the height of residual ridge was more pronounced at the buccal than at the lingual aspect of the extraction socket. Radiographic density of extracted wounds increased by week 4, but the change in density stabilized after week 6. New bone formation was observed at the floor and the peripheral side of extracted socket from week 1. The entrance of extracted socket was sealed by a hard-tissue bridge at week 5. The healing process of extracted wound involved a series of events including new bone formation and residual ridge resorption.

  17. Radiologic study of the healing process of the extracted socket of beagle dogs using cone beam CT

    International Nuclear Information System (INIS)

    Cho, Bong Won; Lee, Won; Lee, Byung Do; Kim, De Sok

    2009-01-01

    To longitudinally observe the healing process of extracted socket and the alterations of the residual ridge in healthy adult dogs using cone beam CT (CBCT). The mandibular premolars of two beagle dogs were removed and the extraction sites were covered with the gingival tissue. CBCTs (3D X-ray CT scanner, Alphard vega, Asahi Co.) were taken at baseline and at 1 week interval for 12 weeks. Radiographic density of extracted wounds was measured on normalized images with a custom-made image analysis program. The amount of alveolar crestal resorption after the teeth extraction was measured with a reformatted three-dimensional image using CBCT. Bony healing pattern of extracted wound of each group was also longitudinally observed and analyzed. Dimensional changes occurred during the first 6 weeks following the extraction of dogs' mandibular premolars. The reduction of the height of residual ridge was more pronounced at the buccal than at the lingual aspect of the extraction socket. Radiographic density of extracted wounds increased by week 4, but the change in density stabilized after week 6. New bone formation was observed at the floor and the peripheral side of extracted socket from week 1. The entrance of extracted socket was sealed by a hard-tissue bridge at week 5. The healing process of extracted wound involved a series of events including new bone formation and residual ridge resorption.

  18. GPU-based streaming architectures for fast cone-beam CT image reconstruction and demons deformable registration

    Energy Technology Data Exchange (ETDEWEB)

    Sharp, G C [Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA 02114 (United States); Kandasamy, N [Electrical and Computer Engineering Department, Drexel University, Philadelphia, PA 19104 (United States); Singh, H [Electrical and Computer Engineering Department, Drexel University, Philadelphia, PA 19104 (United States); Folkert, M [Department of Radiation Oncology, Massachusetts General Hospital, Boston, MA 02114 (United States)

    2007-09-21

    This paper shows how to significantly accelerate cone-beam CT reconstruction and 3D deformable image registration using the stream-processing model. We describe data-parallel designs for the Feldkamp, Davis and Kress (FDK) reconstruction algorithm, and the demons deformable registration algorithm, suitable for use on a commodity graphics processing unit. The streaming versions of these algorithms are implemented using the Brook programming environment and executed on an NVidia 8800 GPU. Performance results using CT data of a preserved swine lung indicate that the GPU-based implementations of the FDK and demons algorithms achieve a substantial speedup-up to 80 times for FDK and 70 times for demons when compared to an optimized reference implementation on a 2.8 GHz Intel processor. In addition, the accuracy of the GPU-based implementations was found to be excellent. Compared with CPU-based implementations, the RMS differences were less than 0.1 Hounsfield unit for reconstruction and less than 0.1 mm for deformable registration.

  19. CT-Guided 125I Seed Interstitial Brachytherapy as a Salvage Treatment for Recurrent Spinal Metastases after External Beam Radiotherapy

    Science.gov (United States)

    Yao, Lihong; Cao, Qianqian; Yang, Jiwen; Meng, Na; Guo, Fuxin; Jiang, Yuliang; Tian, Suqing; Sun, Haitao

    2016-01-01

    The aim of this study is to evaluate the feasibility, safety, and clinical efficacy of CT-guided 125I seed interstitial brachytherapy in patients with recurrent spinal metastases after external beam radiotherapy (EBRT). Between August 2003 and September 2015, 26 spinal metastatic lesions (24 patients) were reirradiated by this salvage therapy modality. Treatment for all patients was preplanned using a three-dimensional treatment planning system 3–5 days before 125I seed interstitial brachytherapy; dosimetry verification was performed immediately after seed implantation. Median actual D 90 was 99 Gy (range, 90–176), and spinal cord median D max was 39 Gy (range, 6–110). Median local control (LC) was 12 months (95% CI: 7.0–17.0). The 6- and 12-month LC rates were 52% and 40%, respectively. Median overall survival (OS) was 11 months (95% CI: 7.7–14.3); 6-month and 1-, 2-, and 3-year OS rates were 65%, 37%, 14%, and 9%, respectively. Pain-free survival ranged from 2 to 42 months (median, 6; 95% CI: 4.6–7.4). Treatment was well-tolerated, with no radiation-induced vertebral compression fractures or myelopathy reported. Reirradiation with CT-guided 125I seed interstitial brachytherapy appears to be feasible, safe, and effective as pain relief or salvage treatment for patients with recurrent spinal metastases after EBRT. PMID:28105434

  20. CT-Guided 125I Seed Interstitial Brachytherapy as a Salvage Treatment for Recurrent Spinal Metastases after External Beam Radiotherapy

    Directory of Open Access Journals (Sweden)

    Lihong Yao

    2016-01-01

    Full Text Available The aim of this study is to evaluate the feasibility, safety, and clinical efficacy of CT-guided 125I seed interstitial brachytherapy in patients with recurrent spinal metastases after external beam radiotherapy (EBRT. Between August 2003 and September 2015, 26 spinal metastatic lesions (24 patients were reirradiated by this salvage therapy modality. Treatment for all patients was preplanned using a three-dimensional treatment planning system 3–5 days before 125I seed interstitial brachytherapy; dosimetry verification was performed immediately after seed implantation. Median actual D90 was 99 Gy (range, 90–176, and spinal cord median Dmax was 39 Gy (range, 6–110. Median local control (LC was 12 months (95% CI: 7.0–17.0. The 6- and 12-month LC rates were 52% and 40%, respectively. Median overall survival (OS was 11 months (95% CI: 7.7–14.3; 6-month and 1-, 2-, and 3-year OS rates were 65%, 37%, 14%, and 9%, respectively. Pain-free survival ranged from 2 to 42 months (median, 6; 95% CI: 4.6–7.4. Treatment was well-tolerated, with no radiation-induced vertebral compression fractures or myelopathy reported. Reirradiation with CT-guided 125I seed interstitial brachytherapy appears to be feasible, safe, and effective as pain relief or salvage treatment for patients with recurrent spinal metastases after EBRT.

  1. A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT

    International Nuclear Information System (INIS)

    Siewerdsen, J.H.; Daly, M.J.; Bakhtiar, B.

    2006-01-01

    X-ray scatter poses a significant limitation to image quality in cone-beam CT (CBCT), resulting in contrast reduction, image artifacts, and lack of CT number accuracy. We report the performance of a simple scatter correction method in which scatter fluence is estimated directly in each projection from pixel values near the edge of the detector behind the collimator leaves. The algorithm operates on the simple assumption that signal in the collimator shadow is attributable to x-ray scatter, and the 2D scatter fluence is estimated by interpolating between pixel values measured along the top and bottom edges of the detector behind the collimator leaves. The resulting scatter fluence estimate is subtracted from each projection to yield an estimate of the primary-only images for CBCT reconstruction. Performance was investigated in phantom experiments on an experimental CBCT benchtop, and the effect on image quality was demonstrated in patient images (head, abdomen, and pelvis sites) obtained on a preclinical system for CBCT-guided radiation therapy. The algorithm provides significant reduction in scatter artifacts without compromise in contrast-to-noise ratio (CNR). For example, in a head phantom, cupping artifact was essentially eliminated, CT number accuracy was restored to within 3%, and CNR (breast-to-water) was improved by up to 50%. Similarly in a body phantom, cupping artifact was reduced by at least a factor of 2 without loss in CNR. Patient images demonstrate significantly increased uniformity, accuracy, and contrast, with an overall improvement in image quality in all sites investigated. Qualitative evaluation illustrates that soft-tissue structures that are otherwise undetectable are clearly delineated in scatter-corrected reconstructions. Since scatter is estimated directly in each projection, the algorithm is robust with respect to system geometry, patient size and heterogeneity, patient motion, etc. Operating without prior information, analytical modeling

  2. 3D printing for orthopedic applications: from high resolution cone beam CT images to life size physical models

    Science.gov (United States)

    Jackson, Amiee; Ray, Lawrence A.; Dangi, Shusil; Ben-Zikri, Yehuda K.; Linte, Cristian A.

    2017-03-01

    With increasing resolution in image acquisition, the project explores capabilities of printing toward faithfully reflecting detail and features depicted in medical images. To improve safety and efficiency of orthopedic surgery and spatial conceptualization in training and education, this project focused on generating virtual models of orthopedic anatomy from clinical quality computed tomography (CT) image datasets and manufacturing life-size physical models of the anatomy using 3D printing tools. Beginning with raw micro CT data, several image segmentation techniques including thresholding, edge recognition, and region-growing algorithms available in packages such as ITK-SNAP, MITK, or Mimics, were utilized to separate bone from surrounding soft tissue. After converting the resulting data to a standard 3D printing format, stereolithography (STL), the STL file was edited using Meshlab, Netfabb, and Meshmixer. The editing process was necessary to ensure a fully connected surface (no loose elements), positive volume with manifold geometry (geometry possible in the 3D physical world), and a single, closed shell. The resulting surface was then imported into a "slicing" software to scale and orient for printing on a Flashforge Creator Pro. In printing, relationships between orientation, print bed volume, model quality, material use and cost, and print time were considered. We generated anatomical models of the hand, elbow, knee, ankle, and foot from both low-dose high-resolution cone-beam CT images acquired using the soon to be released scanner developed by Carestream, as well as scaled models of the skeletal anatomy of the arm and leg, together with life-size models of the hand and foot.

  3. Assessment of image quality in soft tissue and bone visualization tasks for a dedicated extremity cone-beam CT system.

    Science.gov (United States)

    Demehri, S; Muhit, A; Zbijewski, W; Stayman, J W; Yorkston, J; Packard, N; Senn, R; Yang, D; Foos, D; Thawait, G K; Fayad, L M; Chhabra, A; Carrino, J A; Siewerdsen, J H

    2015-06-01

    To assess visualization tasks using cone-beam CT (CBCT) compared to multi-detector CT (MDCT) for musculoskeletal extremity imaging. Ten cadaveric hands and ten knees were examined using a dedicated CBCT prototype and a clinical multi-detector CT using nominal protocols (80 kVp-108mAs for CBCT; 120 kVp- 300 mAs for MDCT). Soft tissue and bone visualization tasks were assessed by four radiologists using five-point satisfaction (for CBCT and MDCT individually) and five-point preference (side-by-side CBCT versus MDCT image quality comparison) rating tests. Ratings were analyzed using Kruskal-Wallis and Wilcoxon signed-rank tests, and observer agreement was assessed using the Kappa-statistic. Knee CBCT images were rated "excellent" or "good" (median scores 5 and 4) for "bone" and "soft tissue" visualization tasks. Hand CBCT images were rated "excellent" or "adequate" (median scores 5 and 3) for "bone" and "soft tissue" visualization tasks. Preference tests rated CBCT equivalent or superior to MDCT for bone visualization and favoured the MDCT for soft tissue visualization tasks. Intraobserver agreement for CBCT satisfaction tests was fair to almost perfect (κ ~ 0.26-0.92), and interobserver agreement was fair to moderate (κ ~ 0.27-0.54). CBCT provided excellent image quality for bone visualization and adequate image quality for soft tissue visualization tasks. • CBCT provided adequate image quality for diagnostic tasks in extremity imaging. • CBCT images were "excellent" for "bone" and "good/adequate" for "soft tissue" visualization tasks. • CBCT image quality was equivalent/superior to MDCT for bone visualization tasks.

  4. Spatial resolution measurements by Radia diagnostic software with SEDENTEXCT image quality phantom in cone beam CT for dental use.

    Science.gov (United States)

    Watanabe, Hiroshi; Nomura, Yoshikazu; Kuribayashi, Ami; Kurabayashi, Tohru

    2018-02-01

    We aimed to employ the Radia diagnostic software with the safety and efficacy of a new emerging dental X-ray modality (SEDENTEXCT) image quality (IQ) phantom in CT, and to evaluate its validity. The SEDENTEXCT IQ phantom and Radia diagnostic software were employed. The phantom was scanned using one medical full-body CT and two dentomaxillofacial cone beam CTs. The obtained images were imported to the Radia software, and the spatial resolution outputs were evaluated. The oversampling method was employed using our original wire phantom as a reference. The resultant modulation transfer function (MTF) curves were compared. The null hypothesis was that MTF curves generated using both methods would be in agreement. One-way analysis of variance tests were applied to the f50 and f10 values from the MTF curves. The f10 values were subjectively confirmed by observing the line pair modules. The Radia software reported the MTF curves on the xy-plane of the CT scans, but could not return f50 and f10 values on the z-axis. The null hypothesis concerning the reported MTF curves on the xy-plane was rejected. There were significant differences between the results of the Radia software and our reference method, except for f10 values in CS9300. These findings were consistent with our line pair observations. We evaluated the validity of the Radia software with the SEDENTEXCT IQ phantom. The data provided were semi-automatic, albeit with problems and statistically different from our reference. We hope the manufacturer will overcome these limitations.

  5. 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT

    NARCIS (Netherlands)

    McDermott, Leah N.; Wendling, Markus; Nijkamp, Jasper; Mans, Anton; Sonke, Jan-Jakob; Mijnheer, Ben J.; van Herk, Marcel

    2008-01-01

    As radiotherapy becomes more complicated, dose and geometry verification become more necessary. The aim of this study was to use back-projected EPID-based 3D in vivo dosimetry and cone-beam CT (CBCT) to obtain a complete account of the entire treatment for a select patient group. Nine

  6. A comparative evaluation of Cone Beam Computed Tomography (CBCT) and Multi-Slice CT (MSCT) Part I. On subjective image quality

    NARCIS (Netherlands)

    Liang, X.; Jacobs, R.; Hassan, B.; Li, L.M.; Pauwels, R.; Corpas, L.; Souza, P.C.; Martens, W.; Alonso, A.; Lambrichts, I.

    2010-01-01

    Aims: To compare image quality and visibility of anatomical structures in the mandible between five Cone Beam Computed Tomography (CBCT) scanners and one Multi-Slice CT (MSCT) system. Materials and methods: One dry mandible was scanned with five CBCT scanners (Accuitomo 3D, i-CAT, NewTom 3G,

  7. Technical Note: Improved CT number stability across patient size using dual-energy CT virtual monoenergetic imaging

    Energy Technology Data Exchange (ETDEWEB)

    Michalak, Gregory; Grimes, Joshua; Fletcher, Joel; Yu, Lifeng; Leng, Shuai; McCollough, Cynthia, E-mail: mccollough.cynthia@mayo.edu [Department of Radiology, Mayo Clinic, Rochester, Minnesota 55905 (United States); Halaweish, Ahmed [Siemens Medical Solutions, Malvern, Pennsylvania 19355 (United States)

    2016-01-15

    Purpose: The purpose of this study was to evaluate, over a wide range of phantom sizes, CT number stability achieved using two techniques for generating dual-energy computed tomography (DECT) virtual monoenergetic images. Methods: Water phantoms ranging in lateral diameter from 15 to 50 cm and containing a CT number test object were scanned on a DSCT scanner using both single-energy (SE) and dual-energy (DE) techniques. The SE tube potentials were 70, 80, 90, 100, 110, 120, 130, 140, and 150 kV; the DE tube potential pairs were 80/140, 70/150Sn, 80/150Sn, 90/150Sn, and 100/150Sn kV (Sn denotes that the 150 kV beam was filtered with a 0.6 mm tin filter). Virtual monoenergetic images at energies ranging from 40 to 140 keV were produced from the DECT data using two algorithms, monoenergetic (mono) and monoenergetic plus (mono+). Particularly in large phantoms, water CT number errors and/or artifacts were observed; thus, datasets with water CT numbers outside ±10 HU or with noticeable artifacts were excluded from the study. CT numbers were measured to determine CT number stability across all phantom sizes. Results: Data exclusions were generally limited to cases when a SE or DE technique with a tube potential of less than 90 kV was used to scan a phantom larger than 30 cm. The 90/150Sn DE technique provided the most accurate water background over the large range of phantom sizes evaluated. Mono and mono+ provided equally improved CT number stability as a function of phantom size compared to SE; the average deviation in CT number was only 1.4% using 40 keV and 1.8% using 70 keV, while SE had an average deviation of 11.8%. Conclusions: The authors’ report demonstrates, across all phantom sizes, the improvement in CT number stability achieved with mono and mono+ relative to SE.

  8. Soft tissue coverage on the segmentation accuracy of the 3D surface-rendered model from cone-beam CT.

    Science.gov (United States)

    Dusseldorp, J K; Stamatakis, H C; Ren, Y

    2017-04-01

    The aim of this study is to investigate the effect of soft tissue presence on the segmentation accuracy of the 3D hard tissue models from cone-beam computed tomography (CBCT). Seven pairs of CBCT Digital Imaging and Communication in Medicine (DICOM) datasets, containing data of human cadaver heads and their respective dry skulls, were used. The effect of the soft tissue presence on the accuracy of the segmented models was evaluated by performing linear and angular measurements and by superimposition and color mapping of the surface discrepancies after splitting the mandible and maxillo-facial complex in the midsagittal plane. The linear and angular measurements showed significant differences for the more posterior transversal measurements on the mandible (p  0.05). The RMSE value for the mandible, however, significantly decreased from 1.240 to 0.981 mm after splitting (p cone-beam CT, below a generally accepted level of clinical significance of 1 mm. However, this level of accuracy may not meet the requirement for applications where high precision is paramount. Accuracy of CBCT-based 3D surface-rendered models, especially of the hard tissues, are crucial in several dental and medical applications, such as implant planning and virtual surgical planning on patients undergoing orthognathic and navigational surgeries. When used in applications where high precision is paramount, the effect of soft tissue presence should be taken into consideration during the segmentation process.

  9. CIAE 600 kV ns pulse neutron generator

    International Nuclear Information System (INIS)

    Shen Guanren; Guan Xialing; Chen Hongtao

    2001-01-01

    The overall composition of CIAE 600 kV ns Pulse Neutron Generator (CPNG) are introduced, and its characteristic, main technological performance and application were also given. CPNG consists of high voltage power supply with highest output voltage 600 kV, direct current 15 mA, stability and ripple ≤0.1%, 2214 mm x 1604 mm x 1504 mm stainless steel high voltage electrode, built in head equipment uniform field accelerating tube, ns pulsed installation, turbomolecular vacuum pump system and drift pipes at 0 degree and 45 degree. Its characteristics are: (1) high current beam; (2) high current beam ns pulsed installation made use of low energy for chopper and high energy for buncher; (3) compactly laid out and simple in structure

  10. CIAE 600 kV ns pulse neutron generator

    CERN Document Server

    Shen Guan Ren; Guan Xia Ling

    2001-01-01

    The overall composition of CIAE 600 kV ns Pulse Neutron Generator (CPNG) are introduced, and its characteristic, main technological performance and application were also given. CPNG consists of high voltage power supply with highest output voltage 600 kV, direct current 15 mA, stability and ripple <=0.1%, 2214 mm x 1604 mm x 1504 mm stainless steel high voltage electrode, built in head equipment uniform field accelerating tube, ns pulsed installation, turbomolecular vacuum pump system and drift pipes at 0 degree and 45 degree. Its characteristics are: (1) high current beam; (2) high current beam ns pulsed installation made use of low energy for chopper and high energy for buncher; (3) compactly laid out and simple in structure

  11. Computed tomography from photon statistics to modern cone-beam CT

    CERN Document Server

    Buzug, T M

    2008-01-01

    Tis book provides an overview of X-ray technology, the historic developmental milestones of modern CT systems, and gives a comprehensive insight into the main reconstruction methods used in computed tomography. Te basis of reconstr- tion is, undoubtedly, mathematics. However, the beauty of computed tomography cannot be understood without a detailed knowledge of X-ray generation, photon- matter interaction, X-ray detection, photon statistics, as well as fundamental signal processing concepts and dedicated measurement systems. Terefore, the reader will ?nd a number of references to these basic d

  12. Distortional effect of beam-hardening artefacts on microCT: a simulation study based on an in vitro caries model.

    Science.gov (United States)

    Kovács, Miklós; Danyi, Róbert; Erdélyi, Miklós; Fejérdy, Pál; Dobó-Nagy, Csaba

    2009-10-01

    The aim of this study was to assess quantitatively the degrading effect of artefacts caused by beam hardening on the microscopic computerized tomography (microCT) measurements of an in vitro caries model. A simulation-based method was described, with which the degrading effect of microCT artefacts on certain parameters of the observed structure could be determined. Simulations were carried out with polychromatic and monochromatic X-ray source, and a linearization method with a second-order polynomial fit algorithm was used in specific cases to correct the beam hardening artefact. The virtual test object was a half-crown of a tooth with an artificial caries lesion. For simulation with monochromatic X-ray source, the relative error of lesion depth and thickness measurements of the remineralized layer was found to be 1%-2%. For polychromatic X-ray source, and omitting beam hardening correction, the relative error exceeded 6%. After appropriate beam-hardening correction, the relative error of the measurement could be reduced to 1%-2%. With the adjustment simulated in this study, microCT having polychromatic X-ray source resulted in the same level of error as with monochromatic source if the linearization method to correct the beam hardening was used. The presented simulation-based method is a useful way to determine artefact-caused distortions for other studies testing objects with different material and geometry.

  13. Dental cone beam ct and its justified use in oral health care

    OpenAIRE

    Jacobs, R

    2011-01-01

    While dental 2D radiology is still the most frequent diagnostic tool, the inherent nature of jaws and teeth might surely benefit from 3D diagnosis. Nowadays, dental cone beam computed tomography may offer high quality images at low radiation doses and costs. Yet, effective dose ranges of CBCT machines may easily vary from 10-1200 micro - sievert, being an equivalent of 2 to 240 dental panoramic radiographs. The same holds true for diagnostic image quality, which exhibits a huge variation amon...

  14. Pin-photodiode array for the measurement of fan-beam energy and air kerma distributions of X-ray CT scanners.

    Science.gov (United States)

    Haba, Tomonobu; Koyama, Shuji; Aoyama, Takahiko; Kinomura, Yutaka; Ida, Yoshihiro; Kobayashi, Masanao; Kameyama, Hiroshi; Tsutsumi, Yoshinori

    2016-07-01

    Patient dose estimation in X-ray computed tomography (CT) is generally performed by Monte Carlo simulation of photon interactions within anthropomorphic or cylindrical phantoms. An accurate Monte Carlo simulation requires an understanding of the effects of the bow-tie filter equipped in a CT scanner, i.e. the change of X-ray energy and air kerma along the fan-beam arc of the CT scanner. To measure the effective energy and air kerma distributions, we devised a pin-photodiode array utilizing eight channels of X-ray sensors arranged at regular intervals along the fan-beam arc of the CT scanner. Each X-ray sensor consisted of two plate type of pin silicon photodiodes in tandem - front and rear photodiodes - and of a lead collimator, which only allowed X-rays to impinge vertically to the silicon surface of the photodiodes. The effective energy of the X-rays was calculated from the ratio of the output voltages of the photodiodes and the dose was calculated from the output voltage of the front photodiode using the energy and dose calibration curves respectively. The pin-photodiode array allowed the calculation of X-ray effective energies and relative doses, at eight points simultaneously along the fan-beam arc of a CT scanner during a single rotation of the scanner. The fan-beam energy and air kerma distributions of CT scanners can be effectively measured using this pin-photodiode array. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  15. An Efficient Estimation Method for Reducing the Axial Intensity Drop in Circular Cone-Beam CT

    Directory of Open Access Journals (Sweden)

    Lei Zhu

    2008-01-01

    Full Text Available Reconstruction algorithms for circular cone-beam (CB scans have been extensively studied in the literature. Since insufficient data are measured, an exact reconstruction is impossible for such a geometry. If the reconstruction algorithm assumes zeros for the missing data, such as the standard FDK algorithm, a major type of resulting CB artifacts is the intensity drop along the axial direction. Many algorithms have been proposed to improve image quality when faced with this problem of data missing; however, development of an effective and computationally efficient algorithm remains a major challenge. In this work, we propose a novel method for estimating the unmeasured data and reducing the intensity drop artifacts. Each CB projection is analyzed in the Radon space via Grangeat's first derivative. Assuming the CB projection is taken from a parallel beam geometry, we extract those data that reside in the unmeasured region of the Radon space. These data are then used as in a parallel beam geometry to calculate a correction term, which is added together with Hu’s correction term to the FDK result to form a final reconstruction. More approximations are then made on the calculation of the additional term, and the final formula is implemented very efficiently. The algorithm performance is evaluated using computer simulations on analytical phantoms. The reconstruction comparison with results using other existing algorithms shows that the proposed algorithm achieves a superior performance on the reduction of axial intensity drop artifacts with a high computation efficiency.

  16. Evaluation of various approaches for assessing dose indicators and patient organ doses resulting from radiotherapy cone-beam CT.

    Science.gov (United States)

    Rampado, Osvaldo; Giglioli, Francesca Romana; Rossetti, Veronica; Fiandra, Christian; Ragona, Riccardo; Ropolo, Roberto

    2016-05-01

    The aim of this study was to evaluate various approaches for assessing patient organ doses resulting from radiotherapy cone-beam CT (CBCT), by the use of thermoluminescent dosimeter (TLD) measurements in anthropomorphic phantoms, a Monte Carlo based dose calculation software, and different dose indicators as presently defined. Dose evaluations were performed on a CBCT Elekta XVI (Elekta, Crawley, UK) for different protocols and anatomical regions. The first part of the study focuses on using pcxmc software (pcxmc 2.0, STUK, Helsinki, Finland) for calculating organ doses, adapting the input parameters to simulate the exposure geometry, and beam dose distribution in an appropriate way. The calculated doses were compared to readouts of TLDs placed in an anthropomorphic Rando phantom. After this validation, the software was used for analyzing organ dose variability associated with patients' differences in size and gender. At the same time, various dose indicators were evaluated: kerma area product (KAP), cumulative air-kerma at the isocenter (Kair), cone-beam dose index, and central cumulative dose. The latter was evaluated in a single phantom and in a stack of three adjacent computed tomography dose index phantoms. Based on the different dose indicators, a set of coefficients was calculated to estimate organ doses for a range of patient morphologies, using their equivalent diameters. Maximum organ doses were about 1 mGy for head and neck and 25 mGy for chest and pelvis protocols. The differences between pcxmc and TLDs doses were generally below 10% for organs within the field of view and approximately 15% for organs at the boundaries of the radiation beam. When considering patient size and gender variability, differences in organ doses up to 40% were observed especially in the pelvic region; for the organs in the thorax, the maximum differences ranged between 20% and 30%. Phantom dose indexes provided better correlation with organ doses than Kair and KAP, with average

  17. Projection correlation based view interpolation for cone beam CT: primary fluence restoration in scatter measurement with a moving beam stop array

    International Nuclear Information System (INIS)

    Yan Hao; Mou Xuanqin; Tang Shaojie; Xu Qiong; Zankl, Maria

    2010-01-01

    Scatter correction is an open problem in x-ray cone beam (CB) CT. The measurement of scatter intensity with a moving beam stop array (BSA) is a promising technique that offers a low patient dose and accurate scatter measurement. However, when restoring the blocked primary fluence behind the BSA, spatial interpolation cannot well restore the high-frequency part, causing streaks in the reconstructed image. To address this problem, we deduce a projection correlation (PC) to utilize the redundancy (over-determined information) in neighbouring CB views. PC indicates that the main high-frequency information is contained in neighbouring angular projections, instead of the current projection itself, which provides a guiding principle that applies to high-frequency information restoration. On this basis, we present the projection correlation based view interpolation (PC-VI) algorithm; that it outperforms the use of only spatial interpolation is validated. The PC-VI based moving BSA method is developed. In this method, PC-VI is employed instead of spatial interpolation, and new moving modes are designed, which greatly improve the performance of the moving BSA method in terms of reliability and practicability. Evaluation is made on a high-resolution voxel-based human phantom realistically including the entire procedure of scatter measurement with a moving BSA, which is simulated by analytical ray-tracing plus Monte Carlo simulation with EGSnrc. With the proposed method, we get visually artefact-free images approaching the ideal correction. Compared with the spatial interpolation based method, the relative mean square error is reduced by a factor of 6.05-15.94 for different slices. PC-VI does well in CB redundancy mining; therefore, it has further potential in CBCT studies.

  18. X-ray absorption microtomography (microCT) and small beam diffraction mapping of sea urchin teeth.

    Science.gov (United States)

    Stock, S R; Barss, J; Dahl, T; Veis, A; Almer, J D

    2002-07-01

    Two noninvasive X-ray techniques, laboratory X-ray absorption microtomography (microCT) and X-ray diffraction mapping, were used to study teeth of the sea urchin Lytechinus variegatus. MicroCT revealed low attenuation regions at near the tooth's stone part and along the carinar process-central prism boundary; this latter observation appears to be novel. The expected variation of Mg fraction x in the mineral phase (calcite, Ca(1-x)Mg(x)CO(3)) cannot account for all of the linear attenuation coefficient decrease in the two zones: this suggested that soft tissue is localized there. Transmission diffraction mapping (synchrotron X-radiation, 80.8 keV, 0.1 x 0.1mm(2) beam area, 0.1mm translation grid, image plate area detector) simultaneously probed variations in 3-D and showed that the crystal elements of the "T"-shaped tooth were very highly aligned. Diffraction patterns from the keel (adaxial web) and from the abaxial flange (containing primary plates and the stone part) differed markedly. The flange contained two populations of identically oriented crystal elements with lattice parameters corresponding to x=0.13 and x=0.32. The keel produced one set of diffraction spots corresponding to the lower x. The compositions were more or less equivalent to those determined by others for camarodont teeth, and the high Mg phase is expected to be disks of secondary mineral epitaxially related to the underlying primary mineral element. Lattice parameter gradients were not noted in the keel or flange. Taken together, the microCT and diffraction results indicated that there was a band of relatively high protein content, of up to approximately 0.25 volume fraction, in the central part of the flange and paralleling its adaxial and abaxial faces. X-ray microCT and microdiffraction data used in conjunction with protein distribution data will be crucial for understanding the properties of various biocomposites and their mechanical functions.

  19. Radiochromic film calibration for the RQT9 quality beam

    Science.gov (United States)

    Costa, K. C.; Gomez, A. M. L.; Alonso, T. C.; Mourao, A. P.

    2017-11-01

    When ionizing radiation interacts with matter it generates energy deposition. Radiation dosimetry is important for medical applications of ionizing radiation due to the increasing demand for diagnostic radiology and radiotherapy. Different dosimetry methods are used and each one has its advantages and disadvantages. The film is a dose measurement method that records the energy deposition by the darkening of its emulsion. Radiochromic films have a little visible light sensitivity and respond better to ionizing radiation exposure. The aim of this study is to obtain the resulting calibration curve by the irradiation of radiochromic film strips, making it possible to relate the darkening of the film with the absorbed dose, in order to measure doses in experiments with X-ray beam of 120 kV, in computed tomography (CT). Film strips of GAFCHROMIC XR-QA2 were exposed according to RQT9 reference radiation, which defines an X-ray beam generated from a voltage of 120 kV. Strips were irradiated in "Laboratório de Calibração de Dosímetros do Centro de Desenvolvimento da Tecnologia Nuclear" (LCD / CDTN) at a dose range of 5-30 mGy, corresponding to the range values commonly used in CT scans. Digital images of the irradiated films were analyzed by using the ImageJ software. The darkening responses on film strips according to the doses were observed and they allowed obtaining the corresponding numeric values to the darkening for each specific dose value. From the numerical values of darkening, a calibration curve was obtained, which correlates the darkening of the film strip with dose values in mGy. The calibration curve equation is a simplified method for obtaining absorbed dose values using digital images of radiochromic films irradiated. With the calibration curve, radiochromic films may be applied on dosimetry in experiments on CT scans using X-ray beam of 120 kV, in order to improve CT acquisition image processes.

  20. A service for monitoring the quality of intraoperative cone beam CT images

    Directory of Open Access Journals (Sweden)

    Heckel Frank

    2016-09-01

    Full Text Available In recent years, operating rooms (ORs have transformed into integrated operating rooms, where devices are able to communicate, exchange data, or even steer and control each other. However, image data processing is commonly done by dedicated workstations for specific clinical use-cases. In this paper, we propose a concept for a dynamic service component for image data processing on the example of automatic image quality assessment (AQUA of intraoperative cone beam computed tomography (CBCT images. The service is build using the Open Surgical Communication Protocol (OSCP and the standard for Digital Imaging and Communications in Medicine (DICOM. We have validated the proposed concept in an integrated demonstrator OR.

  1. A new weighting scheme for cone-beam helical CT to reduce the image noise

    International Nuclear Information System (INIS)

    Taguchi, Katsuyuki; Chiang, B-S S; Silver, Michael D

    2004-01-01

    Reducing the patient dose while keeping the image noise at the same level is desired for x-ray CT examinations. In order to achieve the goal, we propose a new weighting scheme taking the validity of the data and redundant data samples into account. The method is evaluated with a new generalized version of the Feldkamp helical reconstruction algorithm. It allows us to enlarge the projection angular range used in reconstruction, and thus, to reduce the image noise by increasing the detector utilization rate to 100% without sacrificing the image quality or z-resolution. This concept can be adapted to other exact or approximate algorithms as far as they use redundant data samples

  2. SU-F-J-40: Evaluation of Sensitivity of the Automatic Matching Between Cone-Beam CT Image and Simulation CT Image in TrueBeam 2.0 Imaging System 6DoF Considering Different Uncertainty Sources

    Energy Technology Data Exchange (ETDEWEB)

    Bonaque, J; Bautista-Ballesteros, J; Ibanez-Rosello, B; Lliso, F; Carmona, V; Gimeno, J [Hospital La Fe, Valencia, Valencia (Spain); Perez-Calatayud, J [Hospital La Fe, Valencia, Valencia (Spain); Clinica Benidorm, Benidorm, Alicante (Spain)

    2016-06-15

    Purpose: To estimate the sensitivity of TrueBeam 2.0 Imaging System 6DoF automatic matching tool through the acquisition of cone-beam CT images in different phantoms applying submillimeter translations and rotations of tenths of a degree and registered with image simulation CT. Methods: To evaluate overall system-wide image, we consider two uncertainties source; First, the uncertainty of the manual phantom displacement (ε-m). This uncertainty is calculated by a digital caliper (0.01 mm) for vertical (Vrt), lateral (Lat) and longitudinal (Lng). A digital inclinometer (0.01°) for the pitch and roll and the own phantom scale to evaluate the coordinate rotation (Rtn). The second uncertainty is the displacement detected by the algorithm system of matching (σ-d) that we obtain from the standard deviations of the different measurements. We use three different phantoms. The BrainLab Radiosurgery system for supporting masks with an anthropomorphic dummy adapted to allow displacements of 0.1 mm in Vrt, Lat and Lng dimensions and rotations of 0.1° in Pitch dimension. For the analysis of the Rtn and Roll dimensions we use two homemade phantoms (RinoRot and RinoRoll, La Fe Hospital, Valencia, Spain) that allow rotations of 0.3°. Results: In the case of manual displacement of 0.10 ± 0.03 mm in the translations, the system detect 0.10 ± 0.07 mm, 0.12 ± 0.07 mm and 0.13 ± 0.07 mm (mean ± SD) in Lat, Vrt and Lng respectively. In the case of rotational dimension, manual displacement of 0.3 ± 0.1° was detected with 0.19 ± 0.06°, 0.29 ± 0.03° and 0.27 ± 0.06° in Pitch, Roll and Rtn. Conclusion: We conclude that the sensitivity of the automatic matching system is within 0.10 mm in translations and 0.3° in rotations. These values are under the own sensitivity of the software.

  3. Assessment of CT numbers in limited and medium field-of-view scans taken using Accuitomo 170 and Veraviewepocs 3De cone-beam computed tomography scanners

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Matheus L. [Dept. of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, Campinas (Brazil); Tosoni, Guilherme M. [Dept. of Oral Diagnosis and Surgery, Araraquara Dental School, Sao Paulo State University, Araraquara (Brazil); Lindsey, David H.; Mendoza, Kristopher; Tetradis, Sotirios; Mallya, Sanjay M. [Section of Oral and Maxillofacial Radiology, School of Dentistry, University of California, Los Angeles (United States)

    2014-12-15

    To assess the influence of anatomic location on the relationship between computed tomography (CT) number and X-ray attenuation in limited and medium field-of-view (FOV) scans. Tubes containing solutions with different concentrations of K2HPO4 were placed in the tooth sockets of a human head phantom. Cone-beam computed tomography (CBCT) scans were acquired, and CT numbers of the K{sub 2}HPO{sub 4} solutions were measured. The relationship between CT number and K{sub 2}HPO{sub 4} concentration was examined by linear regression analyses. Then, the variation in CT number according to anatomic location was examined. The relationship between K{sub 2}HPO{sub 4} concentration and CT number was strongly linear. The slopes of the linear regressions for the limited FOVs were almost 2-fold lower than those for the medium FOVs. The absolute CT number differed between imaging protocols and anatomic locations. There is a strong linear relationship between X-ray attenuation and CT number. The specific imaging protocol and anatomic location of the object strongly influence this relationship.

  4. Compact 180-kV Marx generator triggered in atmospheric air by femtosecond laser filaments

    Science.gov (United States)

    Arantchouk, L.; Point, G.; Brelet, Y.; Larour, J.; Carbonnel, J.; André, Y.-B.; Mysyrowicz, A.; Houard, A.

    2014-03-01

    We developed a compact Marx generator triggered in atmospheric air by a single femtosecond laser beam undergoing filamentation. Voltage pulses of 180 kV could be generated with a subnanosecond jitter. The same laser beam was also used to initiate simultaneously guided discharges up to 21 cm long at the output of the generator.

  5. GPU accelerated voxel-driven forward projection for iterative reconstruction of cone-beam CT.

    Science.gov (United States)

    Du, Yi; Yu, Gongyi; Xiang, Xincheng; Wang, Xiangang

    2017-01-05

    For cone-beam computed tomography (CBCT), which has been playing an important role in clinical applications, iterative reconstruction algorithms are able to provide advantageous image qualities over the classical FDK. However, the computational speed of iterative reconstruction is a notable issue for CBCT, of which the forward projection calculation is one of the most time-consuming components. In this study, the cone-beam forward projection problem using the voxel-driven model is analysed, and a GPU-based acceleration method for CBCT forward projection is proposed with the method rationale and implementation workflow detailed as well. For method validation and evaluation, computational simulations are performed, and the calculation times of different methods are collected. Compared with the benchmark CPU processing time, the proposed method performs effectively in handling the inter-thread interference problem, and an acceleration ratio as high as more than 100 is achieved compared to a single-threaded CPU implementation. The voxel-driven forward projection calculation for CBCT is highly paralleled by the proposed method, and we believe it will serve as a critical module to develop iterative reconstruction and correction methods for CBCT imaging.

  6. Image quality at low tube voltage (70 kV) and sinogram-affirmed iterative reconstruction for computed tomography in infants with congenital heart disease

    Energy Technology Data Exchange (ETDEWEB)

    Nakagawa, Motoo; Ozawa, Yoshiyuki; Sakurai, Keita; Shimohira, Masashi; Shibamoto, Yuta [Nagoya City University Graduate School of Medical Sciences, Department of Radiology, Nagoya (Japan); Ohashi, Kazuya [Nagoya City University Hospital, Division of Central Radiology, Nagoya (Japan); Asano, Miki [Nagoya City University Graduate School of Medical Sciences, Department of Cardiovascular Surgery, Nagoya (Japan); Yamaguchi, Sachiko [Nagoya City University Graduate School of Medical Sciences, Department of Pediatrics and Neonatology, Nagoya (Japan)

    2015-09-15

    Lower tube voltage has advantages for CT angiography, such as improved contrast To evaluate the image quality of low-voltage (70 kV) CT for congenital heart disease and the ability of sinogram-affirmed iterative reconstruction to improve image quality. Forty-six children with congenital heart disease (median age: 109 days) were examined using dual-source CT. Scans were performed at 80 kV and 70 kV in 21 and 25 children, respectively. A nonionic iodinated contrast medium (300 mg I/ml) was used for the 80-kV protocol. The contrast medium was diluted to 75% (225 mgI/mL) with saline for the 70-kV protocol. Image noise was measured in the two protocols for each group by extracting the standard deviations of a region of interest placed on the descending aorta. We then determined whether sinogram-affirmed iterative reconstruction reduced the image noise at 70 kV. There was more noise at 70 kV than at 80 kV (29 ± 12 vs 20 ± 4.8; P < 0.01). Sinogram-affirmed iterative reconstruction with grade 4 strength settings improved the noise (20 ± 5.9; P < 0.01) for the 70-kV group. Sinogram-affirmed iterative reconstruction improved the image quality of CT in congenital heart disease. (orig.)

  7. Development of a 300-kV Marx generator and its application to drive ...

    Indian Academy of Sciences (India)

    We have indigenously developed a twenty-stage vertical structure type Marx generator. At a matched load of 90 − 100 Ω , for 25 kV DC charging, an output voltage pulse of 230 kV, and duration 150 ns is obtained. This voltage pulse is applied to a relativistic electron beam (REB) planar diode. For a cathode-anode gap of ...

  8. Differential effects of Kv11.1 activators on Kv11.1a, Kv11.1b and Kv11.1a/Kv11.1b channels

    DEFF Research Database (Denmark)

    Larsen, A P; Bentzen, Bo Hjorth; Grunnet, M

    2010-01-01

    K(v)11.1 channels are involved in regulating cellular excitability in various tissues including brain, heart and smooth muscle. In these tissues, at least two isoforms, K(v)11.1a and K(v)11.1b, with different kinetics, are expressed. K(v)11.1 activators are potential therapeutic agents, but their...

  9. Low-dose CT of the paranasal sinuses. Minimizing X-ray exposure with spectral shaping

    Energy Technology Data Exchange (ETDEWEB)

    Wuest, Wolfgang [Friedrich-Alexander-University Erlangen-Nuremberg, Radiological Institute, Erlangen (Germany); Radiological Institute, Erlangen (Germany); May, Matthias; Saake, Marc; Brand, Michael; Uder, Michael; Lell, Michael [Friedrich-Alexander-University Erlangen-Nuremberg, Radiological Institute, Erlangen (Germany)

    2016-11-15

    Shaping the energy spectrum of the X-ray beam has been shown to be beneficial in low-dose CT. This study's aim was to investigate dose and image quality of tin filtration at 100 kV for pre-operative planning in low-dose paranasal CT imaging in a large patient cohort. In a prospective trial, 129 patients were included. 64 patients were randomly assigned to the study protocol (100 kV with additional tin filtration, 150mAs, 192 x 0.6-mm slice collimation) and 65 patients to the standard low-dose protocol (100 kV, 50mAs, 128 x 0.6-mm slice collimation). To assess the image quality, subjective parameters were evaluated using a five-point scale. This scale was applied on overall image quality and contour delineation of critical anatomical structures. All scans were of diagnostic image quality. Bony structures were of good diagnostic image quality in both groups, soft tissues were of sufficient diagnostic image quality in the study group because of a high level of noise. Radiation exposure was very low in both groups, but significantly lower in the study group (CTDI{sub vol} 1.2 mGy vs. 4.4 mGy, p < 0.001). Spectral optimization (tin filtration at 100 kV) allows for visualization of the paranasal sinus with sufficient image quality at a very low radiation exposure. (orig.)

  10. A Feasibility Study of Low-Dose Single-Scan Dual-Energy Cone-Beam CT in Many-View Under-Sampling Framework.

    Science.gov (United States)

    Lee, Donghyeon; Lee, Jiseoc; Kim, Hyoyi; Lee, Taewon; Soh, Jeongtae; Park, Miran; Kim, Changhwan; Lee, Yeon Ju; Cho, Seungryong

    2017-12-01

    A single-scan dual-energy low-dose cone-beam CT (CBCT) imaging technique that exploits a multi-slit filter is proposed in this paper. The multi-slit filter installed between the x-ray source and the scanned object is reciprocated during a scan. The x-ray beams through the slits would generate relatively low-energy x-ray projection data, while the filtered beams would make high-energy projection data. An iterative image reconstruction algorithm that uses an adaptive-steepest-descent method to minimize image total-variation under the constraint of data fidelity was applied to reconstructing the image from the low-energy projection data. Since the high-energy projection data suffer from a substantially high noise level due to the beam filtration, we have developed a new algorithm that exploits the joint sparsity between the low- and high-energy CT images for image reconstruction of the high-energy CT image. The proposed image reconstruction algorithm uses a gradient magnitude image (GMI) of the low-energy CT image by regularizing the difference of GMIs of the low- and high-energy CT images to be minimized. The feasibility of the proposed technique has been demonstrated by the use of various phantoms in the experimental CBCT setup. Furthermore, based on the proposed dual-energy imaging, a material differentiation was performed and its potential utility has been shown. The proposed imaging technique produced promising results for its potential application to a low-dose single-scan dual-energy CBCT.

  11. Scatter Reduction and Correction for Dual-Source Cone-Beam CT Using Prepatient Grids.

    Science.gov (United States)

    Ren, Lei; Chen, Yingxuan; Zhang, You; Giles, William; Jin, Jianyue; Yin, Fang-Fang

    2016-06-01

    Scatter significantly limits the application of the dual-source cone-beam computed tomography by inducing scatter artifacts and degrading contrast-to-noise ratio, Hounsfield-unit accuracy, and image uniformity. Although our previously developed interleaved acquisition mode addressed the cross scatter between the 2 X-ray sources, it doubles the scanning time and doesn't address the forward scatter issue. This study aims to develop a prepatient grid system to address both forward scatter and cross scatter in the dual-source cone-beam computed tomography. Grids attached to both X-ray sources provide physical scatter reduction during the image acquisition. Image data were measured in the unblocked region, while both forward scatter and cross scatter were measured in the blocked region of the projection for postscan scatter correction. Complementary projections were acquired with grids at complementary locations and were merged to form complete projections for reconstruction. Experiments were conducted with different phantom sizes, grid blocking ratios, image acquisition modes, and reconstruction algorithms to investigate their effects on the scatter reduction and correction. The image quality improvement by the prepatient grids was evaluated both qualitatively through the artifact reduction and quantitatively through contrast-to-noise ratio, Hounsfield-unit accuracy, and uniformity using a CATphan 504 phantom. Scatter artifacts were reduced by scatter reduction and were removed by scatter correction method. Contrast-to-noise ratio, Hounsfield-unit accuracy, and image uniformity were improved substantially. The simultaneous acquisition mode achieved comparable contrast-to-noise ratio as the interleaved and sequential modes after scatter reduction and correction. Higher grid blocking ratio and smaller phantom size led to higher contrast-to-noise ratio for the simultaneous mode. The iterative reconstruction with total variation regularization was more effective than the

  12. [Clinical application of cone-beam CT online correcting technology in volume modulated radiation therapy for nasopharyngeal carcinoma].

    Science.gov (United States)

    Qu, Chao; Liang, Guangli; Liu, Guizhi

    2015-07-01

    To investigate the clinical application value of cone-beam CT (CBCT) online correcting technology in volume modulated radiation therapy (VMRT) for nasopharyngeal carcinoma (NPC). Fifty NPC patients applying head neck and shoulder thermoplastic body membrane fixing device were eligible for treatment VMRT, these patients would accept a couple of CBCT scanning by on board imager (OBI) in a fixed time each week after initial setup and after online correcting during the first three week for radiotherapy, CBCT images and DRR images constructed by CT simulation were carried out registration,which could calculate the setup errors of initial setup and after online correcting. Fifty patients were accepted 150 scanning after initial setup and 150 scanning after online correcting respectively, the errors after initial setup were (-1.24 ± 1.25)mm in X direction, (1.19 ± 1.85)mm in Y direction, (1.49 ± 1.70) mm in Z direction. The setup errors after online correcting were (-0.13 ± 0.29)mm in X direction, (0.10 ± 0.47)mm in Y direction, (0.17 ± 0.36)mm in Z direction. The setup errors after online correcting were significantly lower than the errors after initial setup in X direction (P online correcting were 0.46 mm, 0.53 mm and 0.59 mm in X, Y and Z directions respectively. The use of CBCT online correcting technology can significantly reduce the setup errors of VMRT for NPC and improve the treatment effect.

  13. Cone-beam CT with a flat-panel detector: From image science to image-guided surgery

    Science.gov (United States)

    Siewerdsen, Jeffrey H.

    2011-08-01

    The development of large-area flat-panel X-ray detectors (FPDs) has spurred investigation in a spectrum of advanced medical imaging applications, including tomosynthesis and cone-beam CT (CBCT). Recent research has extended image quality metrics and theoretical models to such applications, providing a quantitative foundation for the assessment of imaging performance as well as a general framework for the design, optimization, and translation of such technologies to new applications. For example, cascaded systems models of the Fourier domain metrics, such as noise-equivalent quanta (NEQ), have been extended to these modalities to describe the propagation of signal and noise through the image acquisition and reconstruction chain and to quantify the factors that govern spatial resolution, image noise, and detectability. Moreover, such models have demonstrated basic agreement with human observer performance for a broad range of imaging conditions and imaging tasks. These developments in image science have formed a foundation for the knowledgeable development and translation of CBCT to new applications in image-guided interventions—for example, CBCT implemented on a mobile surgical C-arm for intraoperative 3D imaging. The ability to acquire high-quality 3D images on demand during surgical intervention overcomes conventional limitations of surgical guidance in the context of preoperative images alone. A prototype mobile C-arm developed in academic-industry partnership demonstrates CBCT with low radiation dose, sub-mm spatial resolution, and soft-tissue visibility potentially approaching that of diagnostic CT. Integration of the 3D imaging system with real-time tracking, deformable registration, endoscopic video, and 3D visualization offers a promising addition to the surgical arsenal in interventions ranging from head-and-neck/skull base surgery to spine, orthopaedic, thoracic, and abdominal surgeries. Cadaver studies show the potential for significant boosts in

  14. Tumor Localization Using Cone-Beam CT Reduces Setup Margins in Conventionally Fractionated Radiotherapy for Lung Tumors

    International Nuclear Information System (INIS)

    Yeung, Anamaria R.; Li, Jonathan G.; Shi Wenyin; Newlin, Heather E.; Chvetsov, Alexei; Liu, Chihray; Palta, Jatinder R.; Olivier, Kenneth

    2009-01-01

    Purpose: To determine whether setup margins can be reduced using cone-bea