WorldWideScience

Sample records for automated external defibrillators

  1. Automated External Defibrillator

    Science.gov (United States)

    ... leads to a 10 percent reduction in survival. Training To Use an Automated External Defibrillator Learning how to use an AED and taking a CPR (cardiopulmonary resuscitation) course are helpful. However, if trained ...

  2. Use of Automated External Defibrillators

    Energy Technology Data Exchange (ETDEWEB)

    Gregory K Christensen

    2009-02-01

    In an effort to improve survival from cardiac arrest, the American Heart Association (AHA) has promoted the Chain of Survival concept, describing a sequence of prehospital steps that result in improved survival after sudden cardiac arrest. These interventions include immediate deployment of emergency medical services, prompt cardiopulmonary resuscitation, early defibrillation when indicated, and early initiation of advanced medical care. Early defibrillation has emerged as the most important intervention with survival decreasing by 10% with each minute of delay in defibrillation. Ventricular Fibrillation (VF) is a condition in which there is uncoordinated contraction of the heart cardiac muscle of the ventricles in the heart, making them tremble rather than contract properly. VF is a medical emergency and if the arrhythmia continues for more than a few seconds, blood circulation will cease, and death can occur in a matter of minutes. During VF, contractions of the heart are not synchronized, blood flow ceases, organs begin to fail from oxygen deprivation and within 10 minutes, death will occur. When VF occurs, the victim must be defibrillated in order to establish the heart’s normal rhythm. On average, the wait for an ambulance in populated areas of the United States is about 11 minutes. In view of these facts, the EFCOG Electrical Safety Task Group initiated this review to evaluate the potential value of deployment and use of automated external defibrillators (AEDs) for treatment of SCA victims. This evaluation indicates the long term survival benefit to victims of SCA is high if treated with CPR plus defibrillation within the first 3-5 minutes after collapse. According to the American Heart Association (AHA), survival rates as high as 74% are possible if treatment and defibrillation is performed in the first 3 minutes. In contrast survival rates are only 5% where no AED programs have been established to provide prompt CPR and defibrillation. ["CPR statistics

  3. [ILCOR recommendation on signage of automated external defibrillators (AEDs)].

    Science.gov (United States)

    Truhlár, A

    2010-05-01

    Early defibrillation is a determinant of survival in both out-of-hospital and in-hospital cardiac arrests from ventricular fibrillation and pulseless ventricular tachycardia. The review summarizes importance of early defibrillation with automated external defibrillators (AED) and presents the International Liaison Committee on Resuscitation (ILCOR) recommendation for universal AED sign. The aim of the recommendation is to unify the AED signs worldwide and to spread the knowledge of this. The public in general, but healthcare professionals particularly, should be able to recognize AED location and use the device immediately in case of cardiac arrest.

  4. Automated external defibrillators in the Australian fitness industry.

    Science.gov (United States)

    Norton, Kevin I; Norton, Lynda H

    2008-04-01

    Sudden cardiac arrest (SCA) occurs in many thousands of Australians each year. Scientific evidence shows an increased survival rate for individuals who receive electrical defibrillation in the first few minutes after SCA. In the last decade automated (rhythm-detecting) external defibrillators (AEDs) have become available that are portable and affordable. Although still relatively rare, there is still the potential that SCA may occur when a person undertakes physical activity. Consequently, health/fitness centres are increasingly recognised as higher risk sites that may benefit from placement of AEDs. There are no laws in Australia requiring health/fitness centres to install AEDs. However, several international and professional organisations have "strongly encouraged" larger centres to install AEDs. Guidelines and algorithms are presented to help estimate the risk of SCA in fitness centres. Fitness centre placement is particularly important if the clientele is older or has a 'high-risk' profile, for example, clients with cardiovascular, respiratory or metabolic disease. International negligence case law and duty of care principles suggests the standard of care required in health/fitness centres may be increasing. Therefore, it may be prudent to install AEDs in larger centres and those in which higher risk groups are physically active.

  5. Attitudes toward automated external defibrillator use in Japan.

    Science.gov (United States)

    Taniguchi, Takumi; Omi, Wataru; Inaba, Hideo

    2008-11-01

    The American Heart Association 2005 Guidelines recommend immediate defibrillation for ventricular fibrillation (VF) of short duration, such as witnessed sudden cardiac arrest. However, it is unclear if public-access automated external defibrillators (AEDs) would actually be used in Japan, because there have been few studies about public attitudes regarding AED use. Therefore, we examined Japanese attitudes toward AED use. Between February and March 2006, 3328 individuals, including high school students, teachers, emergency medical technicians (EMTs), medical nurses, and medical students, were asked about AED knowledge and their willingness to operate an AED. All EMTs, 86% of nurses, and 90% of medical students knew how to use AEDs, while only 15% of high school students and 44% of teachers had such knowledge. All EMTs, 78% of nurses, and 94% of medical students reported they would 'definitely' use the AED, but only 12% of high school students and 35% of teachers gave this reply. The reasons for unwillingness to operate AEDs among both laypeople and health care providers were poor of awareness of what AED is and/or how to use an AED. However, 83% of students and 81% of teachers with AED knowledge reported they would 'definitely' use the AED. Many non-medical people in Japan would be unwilling to operate an AED, because they do not know what AED is and/or how to use an AED. However, many would be willing to operate AEDs if they had better understanding of AEDs. Thus, it is necessary to improve public knowledge of AEDs and AED use.

  6. Optimizing a Drone Network to Deliver Automated External Defibrillators.

    Science.gov (United States)

    Boutilier, Justin J; Brooks, Steven C; Janmohamed, Alyf; Byers, Adam; Buick, Jason E; Zhan, Cathy; Schoellig, Angela P; Cheskes, Sheldon; Morrison, Laurie J; Chan, Timothy C Y

    2017-06-20

    Public access defibrillation programs can improve survival after out-of-hospital cardiac arrest, but automated external defibrillators (AEDs) are rarely available for bystander use at the scene. Drones are an emerging technology that can deliver an AED to the scene of an out-of-hospital cardiac arrest for bystander use. We hypothesize that a drone network designed with the aid of a mathematical model combining both optimization and queuing can reduce the time to AED arrival. We applied our model to 53 702 out-of-hospital cardiac arrests that occurred in the 8 regions of the Toronto Regional RescuNET between January 1, 2006, and December 31, 2014. Our primary analysis quantified the drone network size required to deliver an AED 1, 2, or 3 minutes faster than historical median 911 response times for each region independently. A secondary analysis quantified the reduction in drone resources required if RescuNET was treated as a large coordinated region. The region-specific analysis determined that 81 bases and 100 drones would be required to deliver an AED ahead of median 911 response times by 3 minutes. In the most urban region, the 90th percentile of the AED arrival time was reduced by 6 minutes and 43 seconds relative to historical 911 response times in the region. In the most rural region, the 90th percentile was reduced by 10 minutes and 34 seconds. A single coordinated drone network across all regions required 39.5% fewer bases and 30.0% fewer drones to achieve similar AED delivery times. An optimized drone network designed with the aid of a novel mathematical model can substantially reduce the AED delivery time to an out-of-hospital cardiac arrest event. © 2017 American Heart Association, Inc.

  7. Availability of Automated External Defibrillators in Public High Schools.

    Science.gov (United States)

    White, Michelle J; Loccoh, Emefah C; Goble, Monica M; Yu, Sunkyung; Duquette, Deb; Davis, Matthew M; Odetola, Folafoluwa O; Russell, Mark W

    2016-05-01

    To assess automated external defibrillator (AED) distribution and cardiac emergency preparedness in Michigan secondary schools and investigate for association with school sociodemographic characteristics. Surveys were sent via electronic mail to representatives from all public high schools in 30 randomly selected Michigan counties, stratified by population. Association of AED-related factors with school sociodemographic characteristics were evaluated using Wilcoxon rank sum test and χ(2) test, as appropriate. Of 188 schools, 133 (71%) responded to the survey and all had AEDs. Larger student population was associated with fewer AEDs per 100 students (P schools. Schools with >20% students from racial minority groups had significantly fewer AEDs available per 100 students than schools with less racial diversity (P = .03). Schools with more students eligible for free and reduced lunch were less likely to have a cardiac emergency response plan (P = .02) and demonstrated less frequent AED maintenance (P = .03). Although AEDs are available at public high schools across Michigan, the number of AEDs per student varies inversely with minority student population and school size. Unequal distribution of AEDs and lack of cardiac emergency preparedness may contribute to outcomes of sudden cardiac arrest among youth. Copyright © 2016 Elsevier Inc. All rights reserved.

  8. Psychological Effects of Automated External Defibrillator Training A randomized trial

    Science.gov (United States)

    Meischke, Hendrika; Diehr, Paula; Phelps, Randi; Damon, Susan; Rea, Tom

    2011-01-01

    Objectives The objective of this study was to test if an Automated External Defibrillator (AED) training program would positively affect the mental health of family members of high risk patients. Methods 305 ischemic heart disease patients and their family members were randomized to one of four AED training programs: two video-based training programs and two face-to-face training programs that emphasized self-efficacy and perceived control. Patients and family members were surveyed at baseline, 3 and 9 months post ischemic event on demographic characteristics, measures of quality of life (SF=36) , self-efficacy and perceived control. For this study, family members were the focus rather than the patients. Results Regression analyses showed that family members in the face-to-face training programs did not score better on any of the mental health status variables than family members who participated in the other training programs but for an increase in self-efficacy beliefs at 3 months post training. Conclusion The findings suggest that a specifically designed AED training program emphasizing self-efficacy and perceived control beliefs is not likely to enhance family member mental health. PMID:21411144

  9. Use of automated external defibrillators in a Brazilian airline. A 1-year experience

    Directory of Open Access Journals (Sweden)

    Alves Paulo Magalhães

    2001-01-01

    Full Text Available After the incorporation of automated external defibrilators by other airlines and the support of the Brazilian Society of cardiology, Varig Airlines Began the onboard defibrilation program with the initial purpose of equiping wide-body aircrafts frequently used in international flights and that airplanes use in the Rio - São Paulo route. With all fight attendants trained, the automated. External defibrilation devides were incorporated to 34 airplanes of a total pleet of 80 aircrats. The devices were intalled in the bagage compartments secured with velero straps and 2 pairs of electrods, one or which pre-conected to the device to minimize application time. Later, a portable monitor was addres to the ressocitation kit in the long flights. The expansion of the knowledge of the basic life support fundamentors and the correted implantation of the survival chain and of the automated external defibrilators will increase the extense of recovery of cardiorespiratory arrest victins in aircrafts.

  10. Automated external defibrillators in National Collegiate Athletic Association Division I Athletics.

    Science.gov (United States)

    Coris, Eric E; Sahebzamani, Frances; Walz, Steve; Ramirez, Arnold M

    2004-01-01

    Sudden cardiac death is the leading cause of death in athletes. Evidence on current sudden cardiac death prevention through preparticipation history, physicals, and noninvasive cardiovascular diagnostics has demonstrated a low sensitivity for detection of athletes at high risk of sudden cardiac death. Data are lacking on automated external defibrillator programs specifically initiated to respond to rare dysrhythmia in younger, relatively low-risk populations. Surveys were mailed to the head athletic trainers of all National Collegiate Athletic Association Division I athletics programs listed in the National Athletic Trainers' Association directory. In all, 303 surveys were mailed; 186 departments (61%) responded. Seventy-two percent (133) of responding National Collegiate Athletic Association Division I athletics programs have access to automated external defibrillator units; 54% (101) own their units. Proven medical benefit (55%), concern for liability (51%), and affordability (29%) ranked highest in frequency of reasons for automated external defibrillator purchase. Unit cost (odds ratio = 1.01; 95% confidence interval, 1.01-1.0), donated units (odds ratio = 1.92; confidence interval, 3.66-1.01), institution size (odds ratio =.0001; confidence interval, 1.3 E-4 to 2.2E-05), and proven medical benefit of automated external defibrillators (odds ratio = 24; confidence interval, 72-8.1) were the most significant predictors of departmental defibrillator ownership. Emergency medical service response time and sudden cardiac death event history were not significantly predictive of departmental defibrillator ownership. The majority of automated external defibrillator interventions occurred on nonathletes. Many athletics medicine programs are obtaining automated external defibrillators without apparent criteria for determination of need. Usage and maintenance policies vary widely among departments with unit ownership or access. Programs need to approach the issue of unit

  11. The utilization of automated external defibrillators in Taiwan.

    Science.gov (United States)

    Wang, Tsung-Hsi; Wu, Hsi-Wen; Hou, Peter C; Tseng, Hao-Jui

    2018-03-24

    Increasing attention to care of patient succumbed to out-of-hospital cardiac arrest (OHCA) and evidence for improved survival have resulted in many countries to encourage the use automated external defibrillators (AEDs) by legislation. In Taiwan, the amendment of the Emergency Medical Services Act mandated the installation of AEDs in designated areas in 2013. Since then, 6151 AEDs have been installed and registered in mandated and non-mandated locations. The purpose of this study was to investigate the utilization of AEDs at mandated and non-mandated locations. This paper analyzed 217 cases in whom AEDs was used between July 11, 2013 and July 31, 2015. Descriptive statistics were used to analyze the data. The highest frequency of AEDs used was in long-term care facilities, accounting for 34 (15.7%) cases. The second and third highest was in schools and commuting stations. The highest utilization rate of registered AED was in long-term care facilities (73.9%), the second was in residential areas, and the third was in hot spring areas. Employees at the designated locations or medical personnel operated the AED in 143 cases (84.6%), and bystanders, relatives, friends or others operated the AEDs in 26 cases (15.4%). On-site Return of Spontaneous Circulation (ROSC) after applying AEDs occurred in 76 cases (45.8%). Long-term care facilities had the highest utilization of AEDs and government should pay more attention to enforce the installing of AEDs in these places. The government also needs to promote the education public on how to search the AEDs locations. Copyright © 2018. Published by Elsevier B.V.

  12. Wide variation in cardiopulmonary resuscitation interruption intervals among commercially available automated external defibrillators may affect survival despite high defibrillation efficacy.

    Science.gov (United States)

    Snyder, David; Morgan, Carl

    2004-09-01

    Recent studies have associated interruptions of cardiopulmonary resuscitation imposed by automated external defibrillators (AEDs) with poor resuscitation outcome. In particular, the "hands-off" interval between precordial compressions and subsequent defibrillation shock has been implicated. We sought to determine the range of variation among current-generation AEDs with respect to this characteristic. Seven AEDs from six manufacturers were characterized via stopwatch and arrhythmia simulator with respect to the imposed hands-off interval. All AEDs were equipped with new batteries, and measurements were repeated five times for each AED. A wide variation in the hands-off interval between precordial compressions and shock delivery was observed, ranging from 5.2 to 28.4 secs, with only one AED achieving an interruption of <10 secs. Laboratory and clinical data suggest that this range of variation could be responsible for a more than two-fold variation in patient resuscitation success, an effect that far exceeds any defibrillation efficacy differences that may hypothetically exist. In addition to defibrillation waveform and dose, researchers should consider the hands-off cardiopulmonary resuscitation interruption interval between cardiopulmonary resuscitation and subsequent defibrillation shock to be an important covariate of outcome in resuscitation studies. Defibrillator design should minimize this interval to avoid potential adverse consequences on patient survival.

  13. Data management in automated external defibrillators: a call for a standardised solution

    DEFF Research Database (Denmark)

    Nielsen, Anne Møller; Rasmussen, L S

    2011-01-01

    The ECG data stored in automated external defibrillators (AEDs) may be valuable for establishing a final diagnosis and deciding further diagnostics and treatment. Different data management systems are used and this may create significant problems for data storage and access for physicians treating...

  14. Out-of-hospital cardiac arrest: Probability of bystander defibrillation relative to distance to nearest automated external defibrillator.

    Science.gov (United States)

    Sondergaard, Kathrine B; Hansen, Steen Moller; Pallisgaard, Jannik L; Gerds, Thomas Alexander; Wissenberg, Mads; Karlsson, Lena; Lippert, Freddy K; Gislason, Gunnar H; Torp-Pedersen, Christian; Folke, Fredrik

    2018-03-01

    Despite wide dissemination of automated external defibrillators (AEDs), bystander defibrillation rates remain low. We aimed to investigate how route distance to the nearest accessible AED was associated with probability of bystander defibrillation in public and residential locations. We used data from the nationwide Danish Cardiac Arrest Registry and the Danish AED Network to identify out-of-hospital cardiac arrests and route distances to nearest accessible registered AED during 2008-2013. The association between route distance and bystander defibrillation was described using restricted cubic spline logistic regression. We included 6971 out-of-hospital cardiac arrest cases. The proportion of arrests according to distance in meters (≤100, 101-200, >200) to the nearest accessible AED was: 4.6% (n=320), 5.3% (n=370), and 90.1% (n=6281), respectively. For cardiac arrests in public locations, the probability of bystander defibrillation at 0, 100 and 200m from the nearest AED was 35.7% (95% confidence interval 28.0%-43.5%), 21.3% (95% confidence interval 17.4%-25.2%), and 13.7% (95% confidence interval 10.1%-16.8%), respectively. The corresponding numbers for cardiac arrests in residential locations were 7.0% (95% confidence interval -2.1%-16.1%), 1.5% (95% confidence interval 0.002%-2.8%), and 0.9% (95% confidence interval 0.0005%-1.7%), respectively. In public locations, the probability of bystander defibrillation decreased rapidly within the first 100m route distance from cardiac arrest to nearest accessible AED whereas the probability of bystander defibrillation was low for all distances in residential areas. Copyright © 2017 Elsevier B.V. All rights reserved.

  15. Interprofessional education and social interaction: The use of automated external defibrillators in team-based basic life support.

    Science.gov (United States)

    Onan, Arif; Simsek, Nurettin

    2017-04-01

    Automated external defibrillators are pervasive computing devices designed for the treatment and management of acute sudden cardiac arrest. This study aims to explain users' actual use behavior in teams formed by different professions taken after a short time span of interaction with automated external defibrillator. Before the intervention, all the participants were certified with the American Heart Association Basic Life Support for healthcare providers. A statistically significant difference was revealed in mean individual automated external defibrillator technical skills between uniprofessional and interprofessional groups. The technical automated external defibrillator team scores were greater for groups with interprofessional than for those with uniprofessional education. The nontechnical automated external defibrillator skills of interprofessional and uniprofessional teams revealed differences in advantage of interprofessional teams. Students positively accept automated external defibrillators if well-defined and validated training opportunities to use them expertly are available. Uniprofessional teams were successfully supported by their members and, thereby, used automated external defibrillator effectively. Furthermore, the interprofessional approach resulted in as much effective teamwork as the uniprofessional approach.

  16. A qualitative study to identify barriers to deployment and student training in the use of automated external defibrillators in schools

    DEFF Research Database (Denmark)

    Zinckernagel, Line; Hansen, Carolina Malta; Rod, Morten Hulvej

    2017-01-01

    such as delayed access have been reported. The purpose of this study was to identify barriers to the implementation of defibrillator training of students and deployment of defibrillators in schools. Methods: A qualitative study based on semi-structured individual interviews and focus groups with a total of 25......Background: Student training in use of automated external defibrillators and deployment of such defibrillators in schools is recommended to increase survival after out-of-hospital cardiac arrest. Low implementation rates have been observed, and even at schools with a defibrillator, challenges...... to their perception of student training but not for their considerations on the relevance of their placement at schools. Conclusions: It is crucial for implementation of automated external defibrillators in schools to inform staff about how they work and are operated and that students are an appropriate target group...

  17. A qualitative study to identify barriers to deployment and student training in the use of automated external defibrillators in schools

    DEFF Research Database (Denmark)

    Zinckernagel, Line; Hansen, Carolina Malta; Rod, Morten Hulvej

    2017-01-01

    to their perception of student training but not for their considerations on the relevance of their placement at schools. Conclusions: It is crucial for implementation of automated external defibrillators in schools to inform staff about how they work and are operated and that students are an appropriate target group...... for defibrillator training. Furthermore, it is important to provide schools with a basis for decision making about when to install defibrillators, and to ensure that school staff and students are informed about their placement.......Background: Student training in use of automated external defibrillators and deployment of such defibrillators in schools is recommended to increase survival after out-of-hospital cardiac arrest. Low implementation rates have been observed, and even at schools with a defibrillator, challenges...

  18. Utilization of automated external defibrillators installed in commonly used areas of Japanese hospitals

    OpenAIRE

    OHTA, SHOICHI; NAKAO, HIROYUKI; KUSHIMOTO, SHIGEKI; HIRAIDE, ATSUSHI; SAKAMOTO, TETSUYA; NAGAO, KEN; HORI, SHINGO

    2013-01-01

    Objective. Since July 2004, it has become legal in Japan for laypersons to use automated external defibrillators (AEDs). We investigated the effect of AED installation in commonly used areas of Japanese Association for Acute Medicine accredited training (JAAM) hospitals. Methods. In 2008, we sent questionnaires to 419 JAAM hospitals enquiring about the systems, operations, outcome and characteristics of AED usage. Results. Valid responses were received from 271 hospitals (64.7%). A total of 2...

  19. [Worldwide experience with automated external defibrillators: What have we achieved? What else can we expect?].

    Science.gov (United States)

    Trappe, Hans-Joachim

    2016-03-01

    In Germany approximately 70,000-100,000 SCD patients die from sudden cardiac death (SCD). SCD is not caused by a single factor but is a multifactorial problem. In 50 % of SCD victims, sudden cardiac death is the first manifestation of heart disease. SCD is caused by ventricular tachyarrhythmias in approximately 90 % of patients, whereas SCD is caused by bradyarrhythmias in 5-10 % of the patients. Risk stratification is not possible in the majority of them prior to the fatal event. Early defibrillation is the method of choice to terminate ventricular fibrillation. Therefore, it is mandatory to install automatic external defibrillators (AED) in places with many people. There is general agreement that early defibrillation with automated external defibrillators (AED) is an effective tool to treat patients with ventricular fibrillation and will improve survival. It seems necessary to teach cardiocompression and AED use, also to children and adolescents. AED therapy "at home" did not improve survival in patients with cardiac arrest and can not be recommended.

  20. New signal processing algorithms for automated external defibrillators

    OpenAIRE

    Irusta Zarandona, Unai

    2017-01-01

    [ES]La fibrilación ventricular (VF) es el primer ritmo registrado en el 40\\,\\% de las muertes súbitas por paro cardiorrespiratorio extrahospitalario (PCRE). El único tratamiento eficaz para la FV es la desfibrilación mediante una descarga eléctrica. Fuera del hospital, la descarga se administra mediante un desfibrilador externo automático (DEA), que previamente analiza el electrocardiograma (ECG) del paciente y comprueba si presenta un ritmo desfibrilable. La supervivencia en un caso de PCRE ...

  1. Automated external defibrillation as part BLS: implications for education and practice.

    Science.gov (United States)

    Moule, Pam; Albarran, John W

    2002-09-01

    The latest Adult Basic Life Support (BLS) guidelines support the inclusion of the use of the automated external defibrillator (AED), as part of basic life support (BLS). Emphasis on the provision of early defibrillation as part of BLS acknowledges the importance of this manoeuvre in the successful termination of ventricular fibrillation. The ramifications of such changes for both first responders and organisations implementing the guidelines should not be underestimated. Issues relating to resourcing, content and duration of training and retraining, auditing and evaluation require further exploration. To consider these issues now seems particularly pertinent, given the recent launch of the UK Government's paper on public health, 'Saving Lives-Our Healthier Nation' which seeks to deploy AEDs in busy public places for use by trained members of the lay public. Additionally, defibrillation has been identified as one of the key competencies that all trained nurses and other health care providers should be able to undertake. This paper will consider the background to the current guideline changes, analyse the wider implications of translating the recommendations into practice, and offer possible solutions to address the issues raised. Whilst the analysis is particularly pertinent to the United Kingdom, many of the issues raised have international importance.

  2. Impact of Bystander Automated External Defibrillator Use on Survival and Functional Outcomes in Shockable Observed Public Cardiac Arrests.

    Science.gov (United States)

    Pollack, Ross A; Brown, Siobhan P; Rea, Thomas; Aufderheide, Tom; Barbic, David; Buick, Jason E; Christenson, James; Idris, Ahamed H; Jasti, Jamie; Kampp, Michael; Kudenchuk, Peter; May, Susanne; Muhr, Marc; Nichol, Graham; Ornato, Joseph P; Sopko, George; Vaillancourt, Christian; Morrison, Laurie; Weisfeldt, Myron

    2018-02-26

    Background - Survival following out-of-hospital cardiac arrest (OHCA) with shockable rhythms can be improved with early defibrillation. Although shockable OHCA accounts for only ≈25% of overall arrests, ≈60% of public OHCAs are shockable, offering the possibility of restoring thousands of individuals to full recovery with early defibrillation by bystanders. We sought to determine the association of bystander automated external defibrillator use with survival and functional outcomes in shockable observed public OHCA. Methods - From 2011 to 2015, the Resuscitation Outcomes Consortium prospectively collected detailed information on all cardiac arrests at 9 regional centers. The exposures were shock administration by a bystander-applied automated external defibrillator in comparison with initial defibrillation by emergency medical services. The primary outcome measure was discharge with normal or near-normal (favorable) functional status defined as a modified Rankin Score ≤2. Survival to hospital discharge was the secondary outcome measure. Results -Among 49 555 OHCAs, 4115 (8.3%) observed public OHCAs were analyzed, of which 2500 (60.8%) were shockable. A bystander shock was applied in 18.8% of the shockable arrests. Patients shocked by a bystander were significantly more likely to survive to discharge (66.5% versus 43.0%) and be discharged with favorable functional outcome (57.1% versus 32.7%) than patients initially shocked by emergency medical services. After adjusting for known predictors of outcome, the odds ratio associated with a bystander shock was 2.62 (95% confidence interval, 2.07-3.31) for survival to hospital discharge and 2.73 (95% confidence interval, 2.17-3.44) for discharge with favorable functional outcome. The benefit of bystander shock increased progressively as emergency medical services response time became longer. Conclusions - Bystander automated external defibrillator use before emergency medical services arrival in shockable observed

  3. Survival and health care costs until hospital discharge of patients treated with onsite, dispatched or without automated external defibrillator

    NARCIS (Netherlands)

    Berdowski, Jocelyn; Kuiper, Mathijs J.; Dijkgraaf, Marcel G. W.; Tijssen, Jan G. P.; Koster, Rudolph W.

    2010-01-01

    Background: This study aimed to determine whether automated external defibrillator (AED) use during resuscitation is associated with lower in-hospital health care costs. Methods: For this observational prospective study, we included all treated out-of-hospital cardiac arrests of suspected cardiac

  4. Bystander Automated External Defibrillator Use and Clinical Outcomes after Out-of-Hospital Cardiac Arrest

    DEFF Research Database (Denmark)

    Holmberg, Mathias J; Vognsen, Mikael; Andersen, Mikkel S

    2017-01-01

    Aim: To systematically review studies comparing bystander automated external defibrillator (AED) use to no AED use in regard to clinical outcomes in out-of-hospital cardiac arrest (OHCA), and to provide a descriptive summary of studies on the cost-effectiveness of bystander AED use. Methods: We...... randomized trials, and 13 cost-effectiveness studies were included. Meta-analysis of 6 observational studies without critical risk of bias showed that bystander AED use was associated with survival to hospital discharge (all rhythms OR: 1.73 [95% CI: 1.36, 2.18], shockable rhythms OR: 1.66 [95% CI: 1.54, 1.......79]) and favorable neurological outcome (all rhythms OR: 2.12 [95% CI: 1.36, 3.29], shockable rhythms OR: 2.37 [95% CI: 1.58, 3.57]). There was no association between bystander AED use and neurological outcome for non-shockable rhythms (OR: 0.76 [95% CI: 0.10, 5.87]). The Public-Access Defibrillation trial found...

  5. Automated external defibrillators in the hospital: A case of medical reversal.

    Science.gov (United States)

    Stewart, John A

    2018-05-01

    Automated external defibrillators (AEDs) emerged in the 1980s as an important innovation in pre-hospital emergency cardiac care (ECC). In the years since, the American Heart Association (AHA) and the International Liaison Committee for Resuscitation (ILCOR) have promoted AED technology for use in hospitals as well, resulting in the widespread purchase and use of AED-capable defibrillators. In-hospital use of AEDs now appears to have decreased survival from cardiac arrests. This article will look at the use of AEDs in hospitals as a case of "medical reversal." Medical reversal occurs when an accepted, widely used treatment is found to be ineffective or even harmful. This article will discuss the issue of AEDs in the hospital using a conceptual framework provided by recent work on medical reversal. It will go on to consider the implications of the reversal for in-hospital resuscitation programs and emergency medicine more generally. Copyright © 2017 The Author. Published by Elsevier Inc. All rights reserved.

  6. Mobile Versus Fixed Deployment of Automated External Defibrillators in Rural EMS.

    Science.gov (United States)

    Nelson, R Darrell; Bozeman, William; Collins, Greg; Booe, Brian; Baker, Todd; Alson, Roy

    2015-04-01

    There is no consensus on where automated external defibrillators (AEDs) should be placed in rural communities to maximize impact on survival from cardiac arrest. In the community of Stokes County, North Carolina (USA) the Emergency Medical Services (EMS) system promotes cardiopulmonary resuscitation (CPR) public education and AED use with public access defibrillators (PADs) placed mainly in public schools, churches, and government buildings. This study tested the utilization of AEDs assigned to first responders (FRs) in their private-owned-vehicle (POV) compared to AEDs in fixed locations. The authors performed a prospective, observational study measuring utilization of AEDs carried by FRs in their POV compared to utilization of AEDs in fixed locations. Automated external defibrillator utilization is activation with pads placed on the patient and analysis of heart rhythm to determine if shock/no-shock is indicated. The Institutional Review Board of Wake Forest University Baptist Health System approved the study and written informed consent was waived. The study began on December 01, 2012 at midnight and ended on December 01, 2013 at midnight. During the 12-month study period, 81 community AEDs were in place, 66 in fixed locations and 15 assigned to FRs in their POVs. No utilizations of the 66 fixed location AEDs were reported (0.0 utilizations/AED/year) while 19 utilizations occurred in the FR POV AED study group (1.27 utilizations/AED/year; P<.0001). Odds ratio of using a FR POV located AED was 172 times more likely than using a community fixed-location AED in this rural community. Discussion Placing AEDs in a rural community poses many challenges for optimal utilization in terms of cardiac arrest occurrences. Few studies exist to direct rural community efforts in placing AEDs where they can be most effective, and it has been postulated that placing them directly with FRs may be advantageous. In this rural community, the authors found that placing AED devices with

  7. Motivating people to learn cardiopulmonary resuscitation and use of automated external defibrillators.

    Science.gov (United States)

    McDonald, Deborah Dillon; Martin, Deborah; Foley, Diane; Baker, Lee; Hintz, Deborah; Faure, Lauren; Erman, Nancy; Palozie, Jessica; Lundquist, Kathleen; O'Brien, Kara; Prior, Laura; Songco, Narra; Muscillo, Gwyn; Graziani, Denise; Tomczyk, Michael; Price, Sheryl

    2010-01-01

    The purpose of this study was to test the effect of a motivational message on the intention of laypersons to learn cardiopulmonary resuscitation (CPR) and automated external defibrillator (AED) use. A pretest-posttest, double-blind, randomized design was used with 220 community-dwelling adults. Participants were randomly assigned to the treatment group reading the CPR and AED pamphlet emphasizing learning CPR and AED use to save someone they love and the 3-minute window for response time; or to the comparison group reading the identical pamphlet without the 2 motivational statements. Intention to learn CPR and AED use and to look for AEDs in public areas was measured before and after reading the respective pamphlet. No significant difference emerged between the groups for the number of participants planning to learn CPR and AED use. A significant number of participants in both groups increased intention to learn CPR and AED use. Significantly more treatment participants than comparison participants planned to routinely look for AEDs in public areas after reading the pamphlet, however. Teaching critical facts such as the low survival rate for out-of-hospital cardiac arrest might encourage laypersons to learn CPR and AED use. Routinely teaching family members of people at risk for a cardiac arrest about the short window of time in which CPR and AED use must begin and encouraging them to learn about CPR and AEDs to save someone they love may encourage family members to identify the location of AEDs in public places.

  8. Quality of basic life support education and automated external defibrillator setting in schools in Ishikawa, Japan.

    Science.gov (United States)

    Takamura, Akiteru; Ito, Sayori; Maruyama, Kaori; Ryo, Yusuke; Saito, Manami; Fujimura, Shuhei; Ishiura, Yuna; Hori, Ariyuki

    2017-03-01

    Automated external defibrillators (AED) have been installed in schools in Japan since 2004, and the government strongly recommends teaching basic life support (BLS). We therefore examined the quality of BLS education and AED installation in schools. We conducted a prefecture-wide questionnaire survey of all primary and junior high schools in 2016, to assess BLS education and AED installation against the recommendations of the Japan Circulation Society. The results were analyzed using descriptive statistics and chi-squared test. In total, 195 schools out of 315 (62%) responded, of which 38% have introduced BLS education for children. BLS training was held in a smaller proportion of primary schools (18%) than junior high schools (86%). More than 90% of primary school staff had undergone BLS training in the previous 2 years. The most common locations of AED were the gymnasium (32%) followed by entrance hall (28%), staffroom (25%), and infirmary (12%). The reasons given for location were that it was obvious (34%), convenient for staff (32%), could be used out of hours (17%), and the most likely location for a heart attack (15%). Approximately 18% of schools reported that it takes >5 min to reach the AED from the furthest point. BLS training, AED location, and understanding of both are not sufficient to save children's lives efficiently. Authorities should make recommendations about the correct number of AED, and their location, and provide more information to improve the quality of BLS training in schools. © 2016 Japan Pediatric Society.

  9. Preventing Sudden Cardiac Death: Automated External Defibrillators in Ohio High Schools.

    Science.gov (United States)

    Lear, Aaron; Hoang, Minh-Ha; Zyzanski, Stephen J

    2015-10-01

    Ohio passed legislation in 2004 for optional public funding of automated external defibrillators (AEDs) in all Ohio high schools. To report occurrences of sudden cardiac arrest in which AEDs were used in Ohio high schools and to evaluate the adherence of Ohio high schools with AEDs to state law and published guidelines on AEDs and emergency action plans (EAPs) in schools. Cross-sectional survey. Web-based survey. A total of 264 of 827 schools that were members of the Ohio High School Athletic Association. We surveyed schools on AED use, AED maintenance, and EAPs. Twenty-five episodes of AED deployment at 22 schools over an 11-year period were reported; 8 (32%) involved students and 17 (68%) involved adults. The reported survival rate was 60% (n = 15). Most events (n = 20, 80%) in both students and adults occurred at or near athletic facilities. The annual use rate of AEDs was 0.7%. Fifty-three percent (n = 140) of schools reported having an EAP in place for episodes of cardiac arrest. Of the schools with EAPs, 57% (n = 80) reported having rehearsed them. Our data supported the placement of AEDs in high schools given the frequency of use for sudden cardiac arrest and the survival rate reported. They also suggested the need for increased awareness of recommendations for EAPs and the need to formulate and practice EAPs. School EAPs should emphasize planning for events in the vicinity of athletic facilities.

  10. Performance of an automated external defibrillator in a moving ambulance vehicle.

    Science.gov (United States)

    Yun, Jong Geun; Jeung, Kyung Woon; Lee, Byung Kook; Ryu, Hyun Ho; Lee, Hyoung Youn; Kim, Mu Jin; Heo, Tag; Min, Yong Il; You, Yeonho

    2010-04-01

    The available data suggest that automated external defibrillators (AED) can be safely used in vibration-like moving conditions such as rigid inflatable boats and aircraft environments. However, little literature exists examining their performance in a moving ambulance. The present study was undertaken to determine whether an AED is able to analyse the heart rhythm correctly during ambulance transport. An ambulance was driven on paved (20-100 km/h) and unpaved (10 km/h) roads. The performance of two AED devices (CU ER 2, CU Medical Systems Inc., Korea, and Heartstart MRx, Phillips, USA) was determined in a moving ambulance using manikins. Vibration intensity was measured simultaneously with a digital vibrometer. AED performance was then evaluated again on manikins and on a swine model under simulated vibration intensities (0.5-5m/s(2)) measured by the vibrometer in the previous phase of the investigation. The vibration intensity increased with increasing speeds on paved roads (1.98+/-0.44 m/s(2) at 100 km/h). While driving on unpaved roads, it increased to 6.40+/-1.06 m/s(2). Both AED algorithms analysed the heart rhythm correctly under resting state. When tested on pigs, both algorithms showed substantially degraded performances, even at low vibration intensities of 0.5-1m/s(2), which corresponded to vibration intensities while driving on paved roads at 20-60 km/h. This study also showed that electrocardiograms generated on manikins were more resistant to motion artifacts than were the pig electrocardiograms. Ambulance personnel should consider the possibility of misinterpretation by an AED when this device is used while transporting a patient. Copyright (c) 2010 Elsevier Ireland Ltd. All rights reserved.

  11. Presence of automated external defibrillators in North Carolina public middle schools.

    Science.gov (United States)

    Fields, Karl B; Bright, Jacob

    2011-01-01

    Automated external defibrillators (AEDs) have been used in the school setting to successfully resuscitate students, staff, and visitors. All public high schools in North Carolina have an AED. However, the number of North Carolina public middle schools with an AED is unknown. The purpose of this study was to determine the presence of AEDs at public middle schools in North Carolina and to estimate the cost associated with providing an AED to all public middle schools currently without one. All 547 middle schools in North Carolina's 117 public school systems were surveyed in 2009 via e-mail, fax, and, when necessary, telephone about whether an AED was present on site. For middle schools without AEDs, we estimated the cost of purchase and for 1 year of maintenance. A total 66.6% of public middle schools responded to 1 of 3 survey mailings. The remaining schools were contacted by telephone, so that 100% were included in data collection. At the time of the survey, at least 1 AED was present in 334 schools (61.1%). Of the 213 schools without AEDs, 57 (26.8%) were in school systems in which some middle schools had AEDs, and 156 (73.2%) were in systems in which no middle school had an AED. On the basis of a start-up cost of $1,200 per AED, the cost of providing an AED to each school without one is approximately $255,600. These data are based on self-report, and we could not verify whether AEDs were functional. Cost estimates do not include charges for ongoing maintenance and staff training. Two hundred and thirteen North Carolina public middle schools (38.9%) do not have an AED on site.

  12. The use of AEDs by police officers in the City of London. Automated external defibrillators.

    Science.gov (United States)

    Ross, P; Nolan, J; Hill, E; Dawson, J; Whimster, F; Skinner, D

    2001-08-01

    The Guidelines 2000 for cardiopulmonary resuscitation recommend shock delivery to victims in ventricular fibrillation within 5 min of call receipt by the Emergency Medical Services. In an effort to achieve this goal, in some parts of the United States, police officers have been trained to use automated external defibrillators (AEDs). We undertook a 3-year pilot evaluation of the use of AEDs by City of London police (CPOL) officers. Over a period of 3 years, 147 CPOL officers were trained in the use of an AED. Four AEDs were placed on rapid response vehicles covering the City of London. An overall call-response interval target was set at 8 min. The CPOL attended 1103 (90%) of the total of 1232 calls to which they were summoned. The mean interval between the first call received and arrival of the CPOL on scene was 8.9+/-4.0 min. The CPOL applied AEDs to 25 victims, 13 of whom were initially in ventricular fibrillation; at least one shock was delivered to all 13. The interval between call reception and delivery of the first shock was 5.5+/-2.5 min. The mean interval between switching on the AED and delivery of the first shock was 24+/-12 s. Two (15%) of these victims survived to hospital discharge. This study has confirmed the feasibility of training police officers in the UK to use AEDs as first responders. The call received to arrival on scene interval should be reduced by improvements in communication between LAS and CPOL.

  13. Lay bystanders' perspectives on what facilitates cardiopulmonary resuscitation and use of automated external defibrillators in real cardiac arrests

    DEFF Research Database (Denmark)

    Malta Hansen, Carolina; Rosenkranz, Simone Mørk; Folke, Fredrik

    2017-01-01

    is crucial in improving survival, cannot cause substantial harm, and that the AED will provide guidance through CPR; prior hands-on training in AED use; during CPR performance, teamwork (ie, support), using the AED voice prompt and a ventilation mask, as well as demonstrating leadership and feeling a moral......BACKGROUND: Many patients who suffer an out-of-hospital cardiac arrest will fail to receive bystander intervention (cardiopulmonary resuscitation [CPR] or defibrillation) despite widespread CPR training and the dissemination of automated external defibrillators (AEDs). We sought to investigate what......, until data saturation. We used cross-sectional indexing (using software), and inductive in-depth thematic analyses, to identify those factors that facilitated CPR and AED use. In addition to prior hands-on CPR training, the following were described as facilitators: prior knowledge that intervention...

  14. Location of cardiac arrest in a city center: strategic placement of automated external defibrillators in public locations

    DEFF Research Database (Denmark)

    Folke, Fredrik; Lippert, Freddy Knudsen; Nielsen, Søren Loumann

    2009-01-01

    BACKGROUND: Public-access defibrillation with automated external defibrillators (AEDs) is being implemented in many countries worldwide with considerable financial implications. The potential benefit and economic consequences of focused or unfocused AED deployment are unknown. METHODS AND RESULTS......: All cardiac arrests in public in Copenhagen, Denmark, from 1994 through 2005 were geographically located, as were 104 public AEDs placed by local initiatives. In accordance with European Resuscitation Council and American Heart Association (AHA) guidelines, areas with a high incidence of cardiac...... arrests were defined as those with 1 cardiac arrest every 2 or 5 years, respectively. There were 1274 cardiac arrests in public locations. According to the European Resuscitation Council or AHA guidelines, AEDs needed to be deployed in 1.2% and 10.6% of the city area, providing coverage for 19.5% (n=249...

  15. Lay Bystanders' Perspectives on What Facilitates Cardiopulmonary Resuscitation and Use of Automated External Defibrillators in Real Cardiac Arrests

    DEFF Research Database (Denmark)

    Malta Hansen, Carolina; Rosenkranz, Simone Mørk; Folke, Fredrik

    2017-01-01

    BACKGROUND: Many patients who suffer an out-of-hospital cardiac arrest will fail to receive bystander intervention (cardiopulmonary resuscitation [CPR] or defibrillation) despite widespread CPR training and the dissemination of automated external defibrillators (AEDs). We sought to investigate what...... factors encourage lay bystanders to initiate CPR and AED use in a cohort of bystanders previously trained in CPR techniques who were present at an out-of-hospital cardiac arrest. METHODS AND RESULTS: One-hundred and twenty-eight semistructured qualitative interviews with CPR-trained lay bystanders...... to consecutive out-of-hospital cardiac arrest, where an AED was present were conducted (from January 2012 to April 2015, in Denmark). Purposive maximum variation sampling was used to establish the breadth of the bystander perspective. Twenty-six of the 128 interviews were chosen for further in-depth analyses...

  16. Early cardiopulmonary resuscitation and use of Automated External Defibrillators by laypersons in out-of-hospital cardiac arrest using an SMS alert service

    NARCIS (Netherlands)

    Scholten, Annemieke C.; van Manen, Jeanette Gabrielle; van der Worp, Wim E.; IJzerman, Maarten Joost; Doggen, Catharina Jacoba Maria

    2011-01-01

    Aim: To evaluate an SMS service (SMS = short message service = text message) with which laypersons are alerted to go to patients with suspected out-of-hospital cardiac arrest and perform early cardiopulmonary resuscitation (CPR) and use an Automated External Defibrillator (AED). This study is the

  17. Systematic downloading and analysis of data from automated external defibrillators used in out-of-hospital cardiac arrest

    DEFF Research Database (Denmark)

    Hansen, Marco Bo; Lippert, Freddy Knudsen; Rasmussen, Lars Simon

    2014-01-01

    BACKGROUND: Valuable information can be retrieved from automated external defibrillators (AEDs) used in victims of out-of-hospital cardiac arrest (OHCA). We describe our experience with systematic downloading of data from deployed AEDs. The primary aim was to compare the proportion of shockable...... rhythm from AEDs used by laypersons with the corresponding proportion recorded by the Emergency Medical Services (EMS) on arrival. METHODS: In a 20-month study, we collected data on OHCAs in the Capital Region of Denmark where an AED was deployed prior to arrival of EMS. The AEDs were brought...... to the emergency medical dispatch centre for data downloading and rhythm analysis. Patient data were retrieved from the medical records from the admitting hospital, whereas data on EMS rhythm analyses were obtained from the Danish Cardiac Arrest Register between 2001 and 2010. RESULTS: A total of 121 AEDs were...

  18. AEDs at your fingertips: automated external defibrillators on college campuses and a novel approach for increasing accessibility.

    Science.gov (United States)

    Berger, Ryan J; O'Shea, Jesse G

    2014-01-01

    The use of automated external defibrillators (AEDs) increases survival in cardiac arrest events. Due to the success of previous efforts and free, readily available mobile mapping software, the discussion is to emphasize the importance of the use of AEDs to prevent sudden cardiac arrest-related deaths on college campuses and abroad, while suggesting a novel approach to aiding in access and awareness issues. A user-friendly mobile application (a low-cost iOS map) was developed at Florida State University to decrease AED retrieval distance and time. The development of mobile AED maps is feasible for a variety of universities and other entities, with the potential to save lives. Just having AEDs installed is not enough--they need to be easily locatable. Society increasingly relies on phones to provide information, and there are opportunities to use mobile technology to locate and share information about relevant emergency devices; these should be incorporated into the chain of survival.

  19. [Training program on cardiopulmonary resuscitation with the use of automated external defibrillator in a university].

    Science.gov (United States)

    Boaventura, Ana Paula; Miyadahira, Ana Maria Kazue

    2012-03-01

    Early defibrillation in cardiopulmonary resuscitation (CPR) receives increasing emphasis on its priority and rapidity. This is an experience report about the implementation of a training program in CPR using a defibrillator in a private university. The training program in basic CPR maneuvers was based on global guidelines, including a theorical course with practical demonstration of CPR maneuvers with the defibrillator, individual practical training and theoretical and practical assessments. About the performance of students in the practical assessment the mean scores obtained by students in the first stage of the course was 26.4 points, while in the second stage the mean was 252.8 points, in the theoretical assessment the mean in the first stage was 3.06 points and in the second 9.0 points. The implementation of programs like this contribute to the effective acquisition of knowledge (theory) and skill (pratice) for the care of CPR victims.

  20. Neighborhood characteristics, bystander automated external defibrillator use, and patient outcomes in public out-of-hospital cardiac arrest.

    Science.gov (United States)

    Andersen, Lars W; Holmberg, Mathias J; Granfeldt, Asger; Løfgren, Bo; Vellano, Kimberly; McNally, Bryan F; Siegerink, Bob; Kurth, Tobias; Donnino, Michael W

    2018-05-01

    Automated external defibrillators (AEDs) can be used by bystanders to provide rapid defibrillation for patients with out-of-hospital cardiac arrest (OHCA). Whether neighborhood characteristics are associated with AED use is unknown. Furthermore, the association between AED use and outcomes has not been well characterized for all (i.e. shockable and non-shockable) public OHCAs. We included public, non-911-responder witnessed OHCAs registered in the Cardiac Arrest Registry to Enhance Survival (CARES) between 2013 and 2016. The primary patient outcome was survival to hospital discharge with a favorable functional outcome. We first assessed the association between neighborhood characteristics and bystander AED use using logistic regression and then assessed the association between bystander AED use and patient outcomes in a propensity score matched cohort. 25,182 OHCAs were included. Several neighborhood characteristics, including the proportion of people living alone, the proportion of white people, and the proportion with a high-school degree or higher, were associated with bystander AED use. 5132 OHCAs were included in the propensity score-matched cohort. Bystander AED use was associated with an increased risk of a favorable functional outcome (35% vs. 25%, risk difference: 9.7% [95% confidence interval: 7.2%, 12.2%], risk ratio: 1.38 [95% confidence interval: 1.27, 1.50]). This was driven by increased favorable functional outcomes with AED use in patients with shockable rhythms (58% vs. 39%) but not in patients with non-shockable rhythms (10% vs. 10%). Specific neighborhood characteristics were associated with bystander AED use in OHCA. Bystander AED use was associated with an increase in favorable functional outcome. Copyright © 2018 Elsevier B.V. All rights reserved.

  1. The Effect of the Duration of Basic Life Support Training on the Learners' Cardiopulmonary and Automated External Defibrillator Skills

    Science.gov (United States)

    Kang, Ku Hyun; Song, Keun Jeong; Lee, Chang Hee

    2016-01-01

    Background. Basic life support (BLS) training with hands-on practice can improve performance during simulated cardiac arrest, although the optimal duration for BLS training is unknown. This study aimed to assess the effectiveness of various BLS training durations for acquiring cardiopulmonary resuscitation (CPR) and automated external defibrillator (AED) skills. Methods. We randomised 485 South Korean nonmedical college students into four levels of BLS training: level 1 (40 min), level 2 (80 min), level 3 (120 min), and level 4 (180 min). Before and after each level, the participants completed questionnaires regarding their willingness to perform CPR and use AEDs, and their psychomotor skills for CPR and AED use were assessed using a manikin with Skill-Reporter™ software. Results. There were no significant differences between levels 1 and 2, although levels 3 and 4 exhibited significant differences in the proportion of overall adequate chest compressions (p CPR and use AEDs (all, p training provided a moderate level of skill for performing CPR and using AEDs. However, high-quality skills for CPR required longer and hands-on training, particularly hands-on training with AEDs. PMID:27529066

  2. The Effect of the Duration of Basic Life Support Training on the Learners’ Cardiopulmonary and Automated External Defibrillator Skills

    Directory of Open Access Journals (Sweden)

    Jin Hyuck Lee

    2016-01-01

    Full Text Available Background. Basic life support (BLS training with hands-on practice can improve performance during simulated cardiac arrest, although the optimal duration for BLS training is unknown. This study aimed to assess the effectiveness of various BLS training durations for acquiring cardiopulmonary resuscitation (CPR and automated external defibrillator (AED skills. Methods. We randomised 485 South Korean nonmedical college students into four levels of BLS training: level 1 (40 min, level 2 (80 min, level 3 (120 min, and level 4 (180 min. Before and after each level, the participants completed questionnaires regarding their willingness to perform CPR and use AEDs, and their psychomotor skills for CPR and AED use were assessed using a manikin with Skill-Reporter™ software. Results. There were no significant differences between levels 1 and 2, although levels 3 and 4 exhibited significant differences in the proportion of overall adequate chest compressions (p<0.001 and average chest compression depth (p=0.003. All levels exhibited a greater posttest willingness to perform CPR and use AEDs (all, p<0.001. Conclusions. Brief BLS training provided a moderate level of skill for performing CPR and using AEDs. However, high-quality skills for CPR required longer and hands-on training, particularly hands-on training with AEDs.

  3. Acute cardiac events and deployment of emergency medical teams and automated external defibrillators in large football stadiums in the Netherlands.

    Science.gov (United States)

    van de Sandt, Femke; Umans, Victor

    2009-10-01

    The incidence of acute cardiac events - including out-of-hospital cardiac arrest - may be increased in visitors of large sports stadiums when compared with the general population. This study sought to investigate the incidence of acute cardiac events inside large Dutch football stadiums, as well as the emergency response systems deployed in these stadiums and the success rate for in-stadium resuscitation. Retrospective cohort study using a questionnaire sent to the 20 Dutch stadiums that hosted professional matches during the 2006-2007 and 2007-2008 football seasons. Stadium capacity ranged from 3600 to 51 600 spectators. Nearly 13 million spectators attended 686 'Eredivisie' (Honorary Division) and European football matches. All stadiums distribute multiple emergency medical teams among the spectators. Eighty-five percent of the stadiums have an ambulance standby during matches, 95% of the stadiums were equipped with automated external defibrillators (AEDs) during the study period. On an average, one AED was available for every 7576 spectators (range 1800-29 600). Ninety-three cardiac events were reported (7.3 per 1 million spectators). An AED was used 22 times (1.7 per 1 million spectators). Resuscitation was successful in 18 cases (82%, 95% confidence interval: 61-93). The incidence of out-of-hospital cardiac arrest inside large football stadiums in the Netherlands, albeit increased when compared with the general population, is low. The success rate for in-stadium resuscitation by medical teams equipped with AEDs is high. Dutch stadiums appear vigilant in regard to acute cardiac events. This report highlights the importance of adequate emergency medical response systems (including AEDs) in large sports venues.

  4. Efficacy and retention of Basic Life Support education including Automated External Defibrillator usage during a physical education period.

    Science.gov (United States)

    Watanabe, Kae; Lopez-Colon, Dalia; Shuster, Jonathan J; Philip, Joseph

    2017-03-01

    The American Heart Association (AHA) advocates for CPR education as a requirement of secondary school curriculum. Unfortunately, many states have not adopted CPR education. Our aim was to investigate a low-cost, time effective method to educate students on Basic Life Support (BLS), including reeducation. This is a prospective, randomized study. Retention was assessed at 4 months post-initial education. Education was performed by AHA-certified providers during a 45-minute physical education class in a middle school in Florida. This age provides opportunities for reinforcement through high school, with ability for efficient learning. The study included 41 Eighth grade students. Students were randomized into two groups; one group received repeat education 2 months after the first education, the second group did not. All students received BLS education limited to chest compressions and usage of an Automated External Defibrillator. Students had skills and knowledge tests administered pre- and post-education after initial education, and repeated 2 and 4 months later to assess retention. There was a significant increase in CPR skills and knowledge when comparing pre- and post-education results for all time-points ( p  < 0.001). When assessing reeducation, a significant improvement was noted in total knowledge scores but not during the actual steps of CPR. Our study indicates significant increase in CPR knowledge and skills following a one-time 45-minute session. Reeducation may be useful, but the interval needs further investigation. If schools across the United States invested one 45-60-minute period every school year, this would ensure widespread CPR knowledge with minimal cost and loss of school time.

  5. Systematic downloading and analysis of data from automated external defibrillators used in out-of-hospital cardiac arrest.

    Science.gov (United States)

    Hansen, Marco Bo; Lippert, Freddy Knudsen; Rasmussen, Lars Simon; Nielsen, Anne Møller

    2014-12-01

    Valuable information can be retrieved from automated external defibrillators (AEDs) used in victims of out-of-hospital cardiac arrest (OHCA). We describe our experience with systematic downloading of data from deployed AEDs. The primary aim was to compare the proportion of shockable rhythm from AEDs used by laypersons with the corresponding proportion recorded by the Emergency Medical Services (EMS) on arrival. In a 20-month study, we collected data on OHCAs in the Capital Region of Denmark where an AED was deployed prior to arrival of EMS. The AEDs were brought to the emergency medical dispatch centre for data downloading and rhythm analysis. Patient data were retrieved from the medical records from the admitting hospital, whereas data on EMS rhythm analyses were obtained from the Danish Cardiac Arrest Register between 2001 and 2010. A total of 121 AEDs were deployed, of which 91 cases were OHCAs with presumed cardiac origin. The prevalence of initial shockable rhythm was 55.0% (95% CI [44.7-64.8%]). This was significantly greater than the proportion recorded by the EMS (27.6%, 95% CI [27.0-28.3%], p<0.0001). Shockable arrests were significantly more likely to be witnessed (92% vs. 34%, p<0.0001) and the bystander CPR rate was higher (98% vs. 85%, p=0.04). More patients with initial shockable rhythm achieved return of spontaneous circulation upon hospital arrival (88% vs. 7%, p<0.0001) and had higher 30-day survival rate (72% vs. 5%, p<0.0001). AEDs used by laypersons revealed a higher proportion of shockable rhythms compared to the EMS rhythm analyses. Copyright © 2014 The Authors. Published by Elsevier Ireland Ltd.. All rights reserved.

  6. Analysis of bystander CPR quality during out-of-hospital cardiac arrest using data derived from automated external defibrillators.

    Science.gov (United States)

    Fernando, Shannon M; Vaillancourt, Christian; Morrow, Stanley; Stiell, Ian G

    2018-07-01

    Little is known regarding the quality of cardiopulmonary resuscitation (CPR) performed by bystanders in out-of-hospital cardiac arrest (OHCA). We sought to determine quality of bystander CPR provided during OHCA using CPR quality data stored by Automated External Defibrillators (AEDs). We used the Resuscitation Outcomes Consortium database to identify OHCA cases of presumed cardiac etiology where an AED was utilized. We then matched AED data to each case identified. AED data was analyzed using manufacturer software in order to determine overall measures of bystander CPR quality, changes in bystander CPR quality over time, and adherence to existing 2010 Resuscitation Quality Guidelines. 100 cases of OHCA of presumed cardiac etiology involving bystander CPR and with corresponding AED data. Mean age was 62.3 years, and 75% were male. Bystanders demonstrated high-quality CPR over all minutes of resuscitation, with a chest compression fraction of 76%, a compression depth of 5.3 cm, and a compression rate of 111.2 compressions/min. Mean perishock pause was 26.8 s. Adherence rates to 2010 Resuscitation Guidelines for compression rate and depth were found to be 66% and 55%, respectively. CPR quality was lowest in the first minute, resulting from increased delay to rhythm analysis (mean 40.7 s). In cases involving shock delivery, latency from initiation of AED to shock delivery was 59.2 s. We found that bystanders perform high-quality CPR, with strong adherence rates to existing Resuscitation Guidelines. High-quality CPR is maintained over the first five minutes of resuscitation, but was lowest in the first minute. Copyright © 2018 Elsevier B.V. All rights reserved.

  7. Efficacy and retention of Basic Life Support education including Automated External Defibrillator usage during a physical education period

    Directory of Open Access Journals (Sweden)

    Kae Watanabe

    2017-03-01

    Full Text Available The American Heart Association (AHA advocates for CPR education as a requirement of secondary school curriculum. Unfortunately, many states have not adopted CPR education. Our aim was to investigate a low-cost, time effective method to educate students on Basic Life Support (BLS, including reeducation. This is a prospective, randomized study. Retention was assessed at 4 months post-initial education. Education was performed by AHA-certified providers during a 45-minute physical education class in a middle school in Florida. This age provides opportunities for reinforcement through high school, with ability for efficient learning. The study included 41 Eighth grade students. Students were randomized into two groups; one group received repeat education 2 months after the first education, the second group did not. All students received BLS education limited to chest compressions and usage of an Automated External Defibrillator. Students had skills and knowledge tests administered pre- and post-education after initial education, and repeated 2 and 4 months later to assess retention. There was a significant increase in CPR skills and knowledge when comparing pre- and post-education results for all time-points (p < 0.001. When assessing reeducation, a significant improvement was noted in total knowledge scores but not during the actual steps of CPR. Our study indicates significant increase in CPR knowledge and skills following a one-time 45-minute session. Reeducation may be useful, but the interval needs further investigation. If schools across the United States invested one 45–60-minute period every school year, this would ensure widespread CPR knowledge with minimal cost and loss of school time.

  8. Radius of Care in Secondary Schools in the Midwest: Are Automated External Defibrillators Sufficiently Accessible to Enable Optimal Patient Care?

    Science.gov (United States)

    Osterman, Michael; Claiborne, Tina; Liberi, Victor

    2018-04-25

      Sudden cardiac arrest is the leading cause of death among young athletes. According to the American Heart Association, an automated external defibrillator (AED) should be available within a 1- to 1.5-minute brisk walk from the patient for the highest chance of survival. Secondary school personnel have reported a lack of understanding about the proper number and placement of AEDs for optimal patient care.   To determine whether fixed AEDs were located within a 1- to 1.5-minute timeframe from any location on secondary school property (ie, radius of care).   Cross-sectional study.   Public and private secondary schools in northwest Ohio and southeast Michigan.   Thirty schools (24 public, 6 private) volunteered.   Global positioning system coordinates were used to survey the entire school properties and determine AED locations. From each AED location, the radius of care was calculated for 3 retrieval speeds: walking, jogging, and driving a utility vehicle. Data were analyzed to expose any property area that fell outside the radius of care.   Public schools (37.1% ± 11.0%) possessed more property outside the radius of care than did private schools (23.8% ± 8.0%; F 1,28 = 8.35, P = .01). After accounting for retrieval speed, we still observed differences between school types when personnel would need to walk or jog to retrieve an AED ( F 1.48,41.35 = 4.99, P = .02). The percentages of school property outside the radius of care for public and private schools were 72.6% and 56.3%, respectively, when walking and 34.4% and 12.2%, respectively, when jogging. Only 4.2% of the public and none of the private schools had property outside the radius of care when driving a utility vehicle.   Schools should strategically place AEDs to decrease the percentage of property area outside the radius of care. In some cases, placement in a centralized location that is publicly accessible may be more important than the overall number of AEDs on site.

  9. Temporal trends in coverage of historical cardiac arrests using a volunteer-based network of automated external defibrillators accessible to laypersons and emergency dispatch centers.

    Science.gov (United States)

    Hansen, Carolina Malta; Lippert, Freddy Knudsen; Wissenberg, Mads; Weeke, Peter; Zinckernagel, Line; Ruwald, Martin H; Karlsson, Lena; Gislason, Gunnar Hilmar; Nielsen, Søren Loumann; Køber, Lars; Torp-Pedersen, Christian; Folke, Fredrik

    2014-11-18

    Although increased dissemination of automated external defibrillators (AEDs) has been associated with more frequent AED use, the trade-off between the number of deployed AEDs and coverage of cardiac arrests remains unclear. We investigated how volunteer-based AED dissemination affected public cardiac arrest coverage in high- and low-risk areas. All public cardiac arrests (1994-2011) and all registered AEDs (2007-2011) in Copenhagen, Denmark, were identified and geocoded. AED coverage of cardiac arrests was defined as historical arrests ≤100 m from an AED. High-risk areas were defined as those with ≥1 arrest every 2 years and accounted for 1.0% of the total city area. Of 1864 cardiac arrests, 18.0% (n=335) occurred in high-risk areas throughout the study period. From 2007 to 2011, the number of AEDs and the corresponding coverage of cardiac arrests increased from 36 to 552 and from 2.7% to 32.6%, respectively. The corresponding increase for high-risk areas was from 1 to 30 AEDs and coverage from 5.7% to 51.3%, respectively. Since the establishment of the AED network (2007-2011), few arrests (n=55) have occurred ≤100 m from an AED with only 14.5% (n=8) being defibrillated before the arrival of emergency medical services. Despite the lack of a coordinated public access defibrillation program, the number of AEDs increased 15-fold with a corresponding increase in cardiac arrest coverage from 2.7% to 32.6% over a 5-year period. The highest increase in coverage was observed in high-risk areas (from 5.7% to 51.3%). AED networks can be used as useful tools to optimize AED placement in community settings. © 2014 American Heart Association, Inc.

  10. Locating Automated External Defibrillators in a Complicated Urban Environment Considering a Pedestrian-Accessible Network that Focuses on Out-of-Hospital Cardiac Arrests

    Directory of Open Access Journals (Sweden)

    Pil Kwon

    2017-02-01

    Full Text Available Automated external defibrillators (AEDs are portable devices that defibrillate and diagnose sudden-cardiac-arrest patients. Therefore, AEDs are widely installed in public places such as airports, schools, sport complexes, etc., and the installation of AEDs is required by law in these places. However, despite their usefulness, AEDs are mostly installed indoors with limited coverage outdoors. Hence, this study conducts research in the placement of AEDs in outdoor locations. This study considers a complicated urban environment using a pedestrian network dataset and network barriers. We draw on the Teitz and Bart’s (1968 heuristic method that was built in the location-allocation solver in ArcMap. The results of this study found that a total of 455 AEDs, including 227 pre-installed AEDs, could be placed in the study area, thus providing an additional 228 devices. Compared with 10 different installation methods that were set as experimental groups, our test results found that additional installations were able to cover 10% to 30% more actual out-of-hospital cardiac-arrest cases. The main contribution of this study is the proposal of a new method in locating AEDs in optimal areas while considering complicated urban environments. We predict that the cardiac-arrest-related mortality rate would be reduced through implementing the findings of this study.

  11. Impact of community-wide police car deployment of automated external defibrillators on survival from out-of-hospital cardiac arrest.

    Science.gov (United States)

    Myerburg, Robert J; Fenster, Jeffrey; Velez, Mauricio; Rosenberg, Donald; Lai, Shenghan; Kurlansky, Paul; Newton, Starbuck; Knox, Melenda; Castellanos, Agustin

    2002-08-27

    Disappointing survival rates from out-of-hospital cardiac arrests encourage strategies for faster defibrillation, such as use of automated external defibrillators (AEDs) by nonconventional responders. AEDs were provided to all Miami-Dade County, Florida, police. AED-equipped police (P-AED) and conventional emergency medical rescue (EMS) responders are simultaneously deployed to possible cardiac arrests. Times from 9-1-1 contact to the scene were compared for P-AED and concurrently deployed EMS, and both were compared with historical EMS experience. Survival with P-AED was compared with outcomes when EMS was the sole responder. Among 420 paired dispatches of P-AED and EMS, the mean+/-SD P-AED time from 9-1-1 call to arrival at the scene was 6.16+/-4.27 minutes, compared with 7.56+/-3.60 minutes for EMS (P<0.001). Police arrived first to 56% of the calls. The time to first responder arrival among P-AED and EMS was 4.88+/-2.88 minutes (P<0.001), compared with a historical response time of 7.64+/-3.66 minutes when EMS was the sole responder. A 17.2% survival rate was observed for victims with ventricular fibrillation or pulseless ventricular tachycardia (VT/VF), compared with 9.0% for standard EMS before P-AED implementation (P=0.047). However, VT/VF benefit was diluted by the observation that 61% of the initial rhythms were nonshockable, reducing the absolute survival benefit among the total study population to 1.6% (P-AED, 7.6%; EMS, 6.0%). P-AED establishes a layer of responders that generate improved response times and survival from VT/VF. There was no benefit for victims with nonshockable rhythms.

  12. Survival After Ventricular Fibrillation Cardiac Arrest in the Sao Paulo Metropolitan Subway System: First Successful Targeted Automated External Defibrillator (AED) Program in Latin America.

    Science.gov (United States)

    Gianotto-Oliveira, Renan; Gonzalez, Maria Margarita; Vianna, Caio Brito; Monteiro Alves, Maurício; Timerman, Sergio; Kalil Filho, Roberto; Kern, Karl B

    2015-10-09

    Targeted automated external defibrillator (AED) programs have improved survival rates among patients who have an out-of-hospital cardiac arrest (OHCA) in US airports, as well as European and Japanese railways. The Sao Paulo (Brazil) Metro subway carries 4.5 million people per day. A targeted AED program was begun in the Sao Paulo Metro with the objective to improve survival from cardiac arrest. A prospective, longitudinal, observational study of all cardiac arrests in the Sao Paulo Metro was performed from September 2006 through November 2012. This study focused on cardiac arrest by ventricular arrhythmias, and the primary endpoint was survival to hospital discharge with minimal neurological impairment. A total of 62 patients had an initial cardiac rhythm of ventricular fibrillation. Because no data on cardiac arrest treatment or outcomes existed before beginning this project, the first 16 months of the implementation was used as the initial experience and compared with the subsequent 5 years of full operation. Return of spontaneous circulation was not different between the initial 16 months and the subsequent 5 years (6 of 8 [75%] vs. 39 of 54 [72%]; P=0.88). However, survival to discharge was significantly different once the full program was instituted (0 of 8 vs. 23 of 54 [43%]; P=0.001). Implementation of a targeted AED program in the Sao Paulo Metro subway system saved lives. A short interval between arrest and defibrillation was key for good long-term, neurologically intact survival. These results support strategic expansion of targeted AED programs in other large Latin American cities. © 2015 The Authors. Published on behalf of the American Heart Association, Inc., by Wiley Blackwell.

  13. Basic life support and automated external defibrillator skills among ambulance personnel: a manikin study performed in a rural low-volume ambulance setting

    Directory of Open Access Journals (Sweden)

    Nielsen Anne

    2012-05-01

    Full Text Available Abstract Background Ambulance personnel play an essential role in the ‘Chain of Survival’. The prognosis after out-of-hospital cardiac arrest was dismal on a rural Danish island and in this study we assessed the cardiopulmonary resuscitation performance of ambulance personnel on that island. Methods The Basic Life Support (BLS and Automated External Defibrillator (AED skills of the ambulance personnel were tested in a simulated cardiac arrest. Points were given according to a scoring sheet. One sample t test was used to analyze the deviation from optimal care according to the 2005 guidelines. After each assessment, individual feedback was given. Results On 3 consecutive days, we assessed the individual EMS teams responding to OHCA on the island. Overall, 70% of the maximal points were achieved. The hands-off ratio was 40%. Correct compression/ventilation ratio (30:2 was used by 80%. A mean compression depth of 40–50 mm was achieved by 55% and the mean compression depth was 42 mm (SD 7 mm. The mean compression rate was 123 per min (SD 15/min. The mean tidal volume was 746 ml (SD 221 ml. Only the mean tidal volume deviated significantly from the recommended (p = 0.01. During the rhythm analysis, 65% did not perform any visual or verbal safety check. Conclusion The EMS providers achieved 70% of the maximal points. Tidal volumes were larger than recommended when mask ventilation was applied. Chest compression depth was optimally performed by 55% of the staff. Defibrillation safety checks were not performed in 65% of EMS providers.

  14. Optimal Installation Locations for Automated External Defibrillators in Taipei 7-Eleven Stores: Using GIS and a Genetic Algorithm with a New Stirring Operator

    Directory of Open Access Journals (Sweden)

    Chung-Yuan Huang

    2014-01-01

    Full Text Available Immediate treatment with an automated external defibrillator (AED increases out-of-hospital cardiac arrest (OHCA patient survival potential. While considerable attention has been given to determining optimal public AED locations, spatial and temporal factors such as time of day and distance from emergency medical services (EMSs are understudied. Here we describe a geocomputational genetic algorithm with a new stirring operator (GANSO that considers spatial and temporal cardiac arrest occurrence factors when assessing the feasibility of using Taipei 7-Eleven stores as installation locations for AEDs. Our model is based on two AED conveyance modes, walking/running and driving, involving service distances of 100 and 300 meters, respectively. Our results suggest different AED allocation strategies involving convenience stores in urban settings. In commercial areas, such installations can compensate for temporal gaps in EMS locations when responding to nighttime OHCA incidents. In residential areas, store installations can compensate for long distances from fire stations, where AEDs are currently held in Taipei.

  15. Effectiveness of teaching automated external defibrillators use using a traditional classroom instruction versus self-instruction video in non-critical care nurses

    Directory of Open Access Journals (Sweden)

    Ismail M. Saiboon

    2016-04-01

    Full Text Available Objectives: To evaluate the effectiveness and retention of learning automated external defibrillator (AED usage taught through a traditional classroom instruction (TCI method versus a novel self instructed video (SIV technique in non-critical care nurses (NCCN. Methods: A prospective single-blind randomized study was conducted over 7 months (April-October 2014 at the Universiti Kebangsaan Malaysia Medical Center, Kuala Lampur, Malaysia. Eighty nurses were randomized into either TCI or SIV instructional techniques. We assessed knowledge, skill and confidence level at baseline, immediate and 6-months post-intervention. Knowledge and confidence were assessed via questionnaire; skill was assessed by a calibrated and blinded independent assessor using an objective structured clinical examination (OSCE method. Results: Pre-test mean scores for knowledge in the TCI group was 10.87 ± 2.34, and for the SIV group was 10.37 ± 1.85 (maximum achievable score 20.00; 4.05 ± 2.87 in the TCI and 3.71 ± 2.66 in the SIV (maximum score 11.00 in the OSCE evaluation and 9.54 ± 3.65 in the TCI and 8.56 ± 3.47 in the SIV (maximum score 25.00 in the individual’s personal confidence level. Both methods increased the mean scores significantly during immediate post-intervention (0-month. At 6-months, the TCI group scored lower than the SIV group in all aspects 11.13 ± 2.70 versus 12.95 ± 2.26 (p=0.03 in knowledge, 7.27 ± 1.62 versus 7.68 ± 1.73 (p=0.47 in the OSCE, and 16.40 ± 2.72 versus 18.82 ± 3.40 (p=0.03 in confidence level. Conclusion: In NCCN’s, SIV is as good as TCI in providing the knowledge, competency, and confidence in performing AED defibrillation.

  16. A multiple linear regression analysis of factors affecting the simulated Basic Life Support (BLS) performance with Automated External Defibrillator (AED) in Flemish lifeguards.

    Science.gov (United States)

    Iserbyt, Peter; Schouppe, Gilles; Charlier, Nathalie

    2015-04-01

    Research investigating lifeguards' performance of Basic Life Support (BLS) with Automated External Defibrillator (AED) is limited. Assessing simulated BLS/AED performance in Flemish lifeguards and identifying factors affecting this performance. Six hundred and sixteen (217 female and 399 male) certified Flemish lifeguards (aged 16-71 years) performed BLS with an AED on a Laerdal ResusciAnne manikin simulating an adult victim of drowning. Stepwise multiple linear regression analysis was conducted with BLS/AED performance as outcome variable and demographic data as explanatory variables. Mean BLS/AED performance for all lifeguards was 66.5%. Compression rate and depth adhered closely to ERC 2010 guidelines. Ventilation volume and flow rate exceeded the guidelines. A significant regression model, F(6, 415)=25.61, p<.001, ES=.38, explained 27% of the variance in BLS performance (R2=.27). Significant predictors were age (beta=-.31, p<.001), years of certification (beta=-.41, p<.001), time on duty per year (beta=-.25, p<.001), practising BLS skills (beta=.11, p=.011), and being a professional lifeguard (beta=-.13, p=.029). 71% of lifeguards reported not practising BLS/AED. Being young, recently certified, few days of employment per year, practising BLS skills and not being a professional lifeguard are factors associated with higher BLS/AED performance. Measures should be taken to prevent BLS/AED performances from decaying with age and longer certification. Refresher courses could include a formal skills test and lifeguards should be encouraged to practise their BLS/AED skills. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  17. A Smartphone Application to Reduce the Time to Automated External Defibrillator Delivery After a Witnessed Out-of-Hospital Cardiac Arrest: A Randomized Simulation-Based Study.

    Science.gov (United States)

    Hatakeyama, Toshihiro; Nishiyama, Chika; Shimamoto, Tomonari; Kiyohara, Kosuke; Kiguchi, Takeyuki; Chida, Izumi; Izawa, Junichi; Matsuyama, Tasuku; Kitamura, Tetsuhisa; Kawamura, Takashi; Iwami, Taku

    2018-04-13

    We developed a new smartphone application to deliver an automated external defibrillator (AED) to out-of-hospital cardiac arrest scene. The aim of this study was to evaluate whether an AED could be delivered earlier with or without an application in a simulated randomized controlled trial. Participants, who were asked to work as bystanders, were randomly assigned to either an application group or control group and were asked to bring an AED in both groups. The bystanders in the application group sent a signal notification using the application to two responders, who were stationed within 200 meters of the arrest scene, to carry an AED. The primary outcome was the AED delivery time by either the bystander or his/her responder. In total, 61 bystanders were eligible and randomized to either the application group (32) or the control group (29). The 52 with time data were available and analyzed. The AED delivery time by either the bystander or his/her responder was significantly shorter in the application group than in the control group [133.6 (44.4) seconds vs. 202.2 (122.2) seconds, P = 0.01]. In this simulation-based trial, AED delivery time was shortened by our newly developed smartphone application for the bystander to ask nearby responders to find and bring an AED to the cardiac arrest scene (UMIN-Clinical Trials Registry 000016506).This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-No Derivatives License 4.0 (CCBY-NC-ND), where it is permissible to download and share the work provided it is properly cited. The work cannot be changed in any way or used commercially without permission from the journal.

  18. Teaching basic life support with an automated external defibrillator using the two-stage or the four-stage teaching technique.

    Science.gov (United States)

    Bjørnshave, Katrine; Krogh, Lise Q; Hansen, Svend B; Nebsbjerg, Mette A; Thim, Troels; Løfgren, Bo

    2018-02-01

    Laypersons often hesitate to perform basic life support (BLS) and use an automated external defibrillator (AED) because of self-perceived lack of knowledge and skills. Training may reduce the barrier to intervene. Reduced training time and costs may allow training of more laypersons. The aim of this study was to compare BLS/AED skills' acquisition and self-evaluated BLS/AED skills after instructor-led training with a two-stage versus a four-stage teaching technique. Laypersons were randomized to either two-stage or four-stage teaching technique courses. Immediately after training, the participants were tested in a simulated cardiac arrest scenario to assess their BLS/AED skills. Skills were assessed using the European Resuscitation Council BLS/AED assessment form. The primary endpoint was passing the test (17 of 17 skills adequately performed). A prespecified noninferiority margin of 20% was used. The two-stage teaching technique (n=72, pass rate 57%) was noninferior to the four-stage technique (n=70, pass rate 59%), with a difference in pass rates of -2%; 95% confidence interval: -18 to 15%. Neither were there significant differences between the two-stage and four-stage groups in the chest compression rate (114±12 vs. 115±14/min), chest compression depth (47±9 vs. 48±9 mm) and number of sufficient rescue breaths between compression cycles (1.7±0.5 vs. 1.6±0.7). In both groups, all participants believed that their training had improved their skills. Teaching laypersons BLS/AED using the two-stage teaching technique was noninferior to the four-stage teaching technique, although the pass rate was -2% (95% confidence interval: -18 to 15%) lower with the two-stage teaching technique.

  19. Effects of introducing a voluntary virtual patient module to a basic life support with an automated external defibrillator course: a randomised trial

    Directory of Open Access Journals (Sweden)

    Kononowicz Andrzej A

    2012-06-01

    Full Text Available Abstract Background The concept of virtual patients (VPs encompasses a great variety of predominantly case-based e-learning modules with different complexity and fidelity levels. Methods for effective placement of VPs in the process of medical education are sought. The aim of this study was to determine whether the introduction of a voluntary virtual patients module into a basic life support with an automated external defibrillator (BLS-AED course improved the knowledge and skills of students taking the course. Methods Half of the students were randomly assigned to an experimental group and given voluntary access to a virtual patient module consisting of six cases presenting BLS-AED knowledge and skills. Pre- and post-course knowledge tests and skills assessments were performed, as well as a survey of students' satisfaction with the VP usage. In addition, time spent using the virtual patient system, percentage of screen cards viewed and scores in the formative questions in the VP system throughout the course were traced and recorded. Results The study was conducted over a six week period and involved 226 first year medical students. The voluntary module was used by 61 (54% of the 114 entitled study participants. The group that used VPs demonstrated better results in knowledge acquisition and in some key BLS-AED action skills than the group without access, or those students from the experimental group deliberately not using virtual patients. Most of the students rated the combination of VPs and corresponding teaching events positively. Conclusions The overall positive reaction of students and encouraging results in knowledge and skills acquisition suggest that the usage of virtual patients in a BLS-AED course on a voluntary basis is feasible and should be further investigated.

  20. Effects of introducing a voluntary virtual patient module to a basic life support with an automated external defibrillator course: a randomised trial.

    Science.gov (United States)

    Kononowicz, Andrzej A; Krawczyk, Paweł; Cebula, Grzegorz; Dembkowska, Marta; Drab, Edyta; Frączek, Bartosz; Stachoń, Aleksandra J; Andres, Janusz

    2012-06-18

    The concept of virtual patients (VPs) encompasses a great variety of predominantly case-based e-learning modules with different complexity and fidelity levels. Methods for effective placement of VPs in the process of medical education are sought. The aim of this study was to determine whether the introduction of a voluntary virtual patients module into a basic life support with an automated external defibrillator (BLS-AED) course improved the knowledge and skills of students taking the course. Half of the students were randomly assigned to an experimental group and given voluntary access to a virtual patient module consisting of six cases presenting BLS-AED knowledge and skills. Pre- and post-course knowledge tests and skills assessments were performed, as well as a survey of students' satisfaction with the VP usage. In addition, time spent using the virtual patient system, percentage of screen cards viewed and scores in the formative questions in the VP system throughout the course were traced and recorded. The study was conducted over a six week period and involved 226 first year medical students. The voluntary module was used by 61 (54%) of the 114 entitled study participants. The group that used VPs demonstrated better results in knowledge acquisition and in some key BLS-AED action skills than the group without access, or those students from the experimental group deliberately not using virtual patients. Most of the students rated the combination of VPs and corresponding teaching events positively. The overall positive reaction of students and encouraging results in knowledge and skills acquisition suggest that the usage of virtual patients in a BLS-AED course on a voluntary basis is feasible and should be further investigated.

  1. [Research and Design of a System for Detecting Automated External Defbrillator Performance Parameters].

    Science.gov (United States)

    Wang, Kewu; Xiao, Shengxiang; Jiang, Lina; Hu, Jingkai

    2017-09-30

    In order to regularly detect the performance parameters of automated external defibrillator (AED), to make sure it is safe before using the instrument, research and design of a system for detecting automated external defibrillator performance parameters. According to the research of the characteristics of its performance parameters, combing the STM32's stability and high speed with PWM modulation control, the system produces a variety of ECG normal and abnormal signals through the digital sampling methods. Completed the design of the hardware and software, formed a prototype. This system can accurate detect automated external defibrillator discharge energy, synchronous defibrillation time, charging time and other key performance parameters.

  2. [Public access defibrillation: successful cardiopulmonary resuscitation due to automatic external defibrillator at traffic accident].

    Science.gov (United States)

    Wanke, S; Reuter, H; Pfister, R; Michels, G

    2014-03-01

    A 65-year-old man collapsed after he stepped out of his car after a traffic accident. Fortunately, two police officers on a routine patrol in the area were quickly on the scene and started cardiopulmonary resuscitation. A passerby noticed that the patient was in distress and that an automatic defibrillator was nearby. He attached the electrodes of the defibrillator to the chest of the patient in accordance with instructions on the defibrillator and terminated the ventricular fibrillation (200 joule, biphasic). Emergency cardiac catheterization revealed a subtotal stenosis proximally in the right coronary artery, which was successfully treated with a stent. Based on the ideal basic life support, the immediate care by emergency mobile system and coronary angioplasty with successful revascularisation the patient could be released without any neurological deficit. This case illustrates that laypersons can use automatic external defibrillator in case of cardiac resuscitation sufficiently and quickly. © Georg Thieme Verlag KG Stuttgart · New York.

  3. Providers with Limited Experience Perform Better in Advanced Life Support with Assistance Using an Interactive Device with an Automated External Defibrillator Linked to a Ventilator.

    Science.gov (United States)

    Busch, Christian Werner; Qalanawi, Mohammed; Kersten, Jan Felix; Kalwa, Tobias Johannes; Scotti, Norman Alexander; Reip, Wikhart; Doehn, Christoph; Maisch, Stefan; Nitzschke, Rainer

    2015-10-01

    Medical teams with limited experience in performing advanced life support (ALS) or with a low frequency of cardiopulmonary resuscitation (CPR) while on duty, often have difficulty complying with CPR guidelines. This study evaluated whether the quality of CPR of trained medical students, who served as an example of teams with limited experience in ALS, could be improved with device assistance. The primary outcome was the hands-off time (i.e., the percentage of the entire CPR time without chest compressions). The secondary outcome was seven time intervals, which should be as short as possible, and the quality of ventilations and chest compressions on the mannequin. We compared standard CPR equipment to an interactive device with visual and acoustic instructions for ALS workflow measures to guide briefly trained medical students through the ALS algorithm in a full-scale mannequin simulation study with a randomized crossover study design. The study equipment consisted of an automatic external defibrillator and ventilator that were electronically linked and communicating as a single system. Included were regular medical students in the third to sixth years of medical school of one class who provided written informed consent for voluntary participation and for the analysis of their CPR performance data. No exclusion criteria were applied. For statistical measures of evaluation we used an analysis of variance for crossover trials accounting for treatment effect, sequence effect, and carry-over effect, with adjustment for prior practical experience of the participants. Forty-two medical students participated in 21 CPR sessions, each using the standard and study equipment. Regarding the primary end point, the study equipment reduced the hands-off time from 40.1% (95% confidence interval [CI] 36.9-43.4%) to 35.6% (95% CI 32.4-38.9%, p = 0.031) compared with the standard equipment. Within the prespecified secondary end points, study equipment reduced the time interval until

  4. "It Doesn't Make Sense for Us Not to Have One"-Understanding Reasons Why Community Sports Organizations Chose to Participate in a Funded Automated External Defibrillator Program.

    Science.gov (United States)

    Fortington, Lauren V; Bekker, Sheree; Morgan, Damian; Finch, Caroline F

    2017-10-10

    Implementation of automated external defibrillators (AEDs) in community sports settings is an important component of emergency medical planning. This study aimed to understand motivations for why sports organizations participated in a government-funded program that provided AEDs and associated first-aid training. Face-to-face interviews. Community sports organizations in Victoria, Australia. Representatives from 14 organizations who participated in a government-funded AED program. Motivations to participate in the AED program were explored using a qualitative descriptive approach. Two overarching themes emerged: awareness of the program and decision to apply. Awareness was gained indirectly through grant advertising in newsletters/emails/web sites and directly through their sporting associations. For most organizations, there was no decision process per se, rather, the opportunity to apply was the key determinant for participating in the program. A duty of care also emerged as a key driving factor, with recognition of AEDs as a valuable asset to communities broadly, not just the participants' immediate sports setting. Reflecting on participation in the program, these participants identified that it was important to increase awareness about AED ownership and use. The program benefits were clearly summed up as being best prepared for a worst-case scenario. This study provides new understanding of why community sports organizations apply for an AED and training. The strongest reason was simply the opportunity to acquire this at no cost. Therefore, for wider implementation of AEDs, additional funding opportunities, targeted awareness of these opportunities, and continued promotion of AED importance are recommended.

  5. Use of a geographic information system to identify differences in automated external defibrillator installation in urban areas with similar incidence of public out-of-hospital cardiac arrest: a retrospective registry-based study.

    Science.gov (United States)

    Fredman, David; Haas, Jan; Ban, Yifang; Jonsson, Martin; Svensson, Leif; Djarv, Therese; Hollenberg, Jacob; Nordberg, Per; Ringh, Mattias; Claesson, Andreas

    2017-06-02

    Early defibrillation in out-of-hospital cardiac arrest (OHCA) is of importance to improve survival. In many countries the number of automated external defibrillators (AEDs) is increasing, but the use is low. Guidelines suggest that AEDs should be installed in densely populated areas and in locations with many visitors. Attempts have been made to identify optimal AED locations based on the incidence of OHCA using geographical information systems (GIS), but often on small datasets and the studies are seldom reproduced. The aim of this paper is to investigate if the distribution of public AEDs follows the incident locations of public OHCAs in urban areas of Stockholm County, Sweden. OHCA data were obtained from the Swedish Register for Cardiopulmonary Resuscitation and AED data were obtained from the Swedish AED Register. Urban areas in Stockholm County were objectively classified according to the pan-European digital mapping tool, Urban Atlas (UA). Furthermore, we reclassified and divided the UA land cover data into three classes (residential, non-residential and other areas). GIS software was used to spatially join and relate public AED and OHCA data and perform computations on relations and distance. Between 1 January 2012 and 31 December 2014 a total of 804 OHCAs occurred in public locations in Stockholm County and by December 2013 there were 1828 AEDs available. The incidence of public OHCAs was similar in residential (47.3%) and non-residential areas (43.4%). Fewer AEDs were present in residential areas than in non-residential areas (29.4% vs 68.8%). In residential areas the median distance between OHCAs and AEDs was significantly greater than in non-residential areas (288 m vs 188 m, p<0.001). The majority of public OHCAs occurred in areas classified in UA as 'residential areas' with limited AED accessibility. These areas need to be targeted for AED installation and international guidelines need to take geographical location into account when suggesting

  6. Effect of an interactive cardiopulmonary resuscitation assist device with an automated external defibrillator synchronised with a ventilator on the CPR performance of emergency medical service staff: a randomised simulation study.

    Science.gov (United States)

    Nitzschke, Rainer; Doehn, Christoph; Kersten, Jan F; Blanz, Julian; Kalwa, Tobias J; Scotti, Norman A; Kubitz, Jens C

    2017-04-04

    The present study evaluates whether the quality of advanced cardiac life support (ALS) is improved with an interactive prototype assist device. This device consists of an automated external defibrillator linked to a ventilator and provides synchronised visual and acoustic instructions for guidance through the ALS algorithm and assistance for face-mask ventilations. We compared the cardiopulmonary resuscitation (CPR) quality of emergency medical system (EMS) staff members using the study device or standard equipment in a mannequin simulation study with a prospective, controlled, randomised cross-over study design. Main outcome was the effect of the study device compared to the standard equipment and the effect of the number of prior ALS trainings of the EMS staff on the CPR quality. Data were analysed using analyses of covariance (ANCOVA) and binary logistic regression, accounting for the study design. In 106 simulations of 56 two-person rescuer teams, the mean hands-off time was 24.5% with study equipment and 23.5% with standard equipment (Difference 1.0% (95% CI: -0.4 to 2.5%); p = 0.156). With both types of equipment, the hands-off time decreased with an increasing cumulative number of previous CPR trainings (p = 0.042). The study equipment reduced the mean time until administration of adrenaline (epinephrine) by 23 s (p = 0.003) and that of amiodarone by 17 s (p = 0.016). It also increased the mean number of changes in the person doing chest compressions (0.6 per simulation; p < 0.001) and decreased the mean number of chest compressions (2.8 per minute; p = 0.022) and the mean number of ventilations (1.8 per minute; p < 0.001). The chance of administering amiodarone at the appropriate time was higher, with an odds ratio of 4.15, with the use of the study equipment CPR.com compared to the standard equipment (p = 0.004). With an increasing number of prior CPR trainings, the time intervals in the ALS algorithm until the

  7. Estimation of current density distribution under electrodes for external defibrillation

    Directory of Open Access Journals (Sweden)

    Papazov Sava P

    2002-12-01

    Full Text Available Abstract Background Transthoracic defibrillation is the most common life-saving technique for the restoration of the heart rhythm of cardiac arrest victims. The procedure requires adequate application of large electrodes on the patient chest, to ensure low-resistance electrical contact. The current density distribution under the electrodes is non-uniform, leading to muscle contraction and pain, or risks of burning. The recent introduction of automatic external defibrillators and even wearable defibrillators, presents new demanding requirements for the structure of electrodes. Method and Results Using the pseudo-elliptic differential equation of Laplace type with appropriate boundary conditions and applying finite element method modeling, electrodes of various shapes and structure were studied. The non-uniformity of the current density distribution was shown to be moderately improved by adding a low resistivity layer between the metal and tissue and by a ring around the electrode perimeter. The inclusion of openings in long-term wearable electrodes additionally disturbs the current density profile. However, a number of small-size perforations may result in acceptable current density distribution. Conclusion The current density distribution non-uniformity of circular electrodes is about 30% less than that of square-shaped electrodes. The use of an interface layer of intermediate resistivity, comparable to that of the underlying tissues, and a high-resistivity perimeter ring, can further improve the distribution. The inclusion of skin aeration openings disturbs the current paths, but an appropriate selection of number and size provides a reasonable compromise.

  8. What Is an Automated External Defibrillator?

    Science.gov (United States)

    ... such as police, fire service personnel, flight attendants, security guards and other lay rescuers who have been ... venues, shopping malls, airports, airplanes, businesses, convention centers, hotels, schools and doctors’ offices. They should also be ...

  9. Uso do desfibrilador automático externo no ambiente pré-hospitalar peruano: melhorando a resposta a emergências na América Latina Use of automated external defibrillator in Peruvian out-of-hospital environment: improving emergency response in Latin America

    Directory of Open Access Journals (Sweden)

    Pablo Lister

    2009-08-01

    Full Text Available Este relato de caso reporta o atendimento pré-hospitalar de um paciente com fatores de risco atendido pelo serviço pré-hospitalar ao ser acometido por uma parada cardíaca e apresentar fibrilação ventricular. O paciente foi atendido seguindo os padrões de suporte básico de vida e suporte cardiovascular avançado. Um desfibrilador automático externo (DAE foi aplicado com resultados favoráveis e o paciente se recuperou do quadro de perigo de vida com sucesso. Este é o primeiro relato documentado com resultados favoráveis no Peru, na área de atendimento pré-hospitalar e enfatiza a necessidade de serem adotadas políticas de acesso público à desfibrilação precoce.El presente reporte de caso, relata la atencion prehospitalaria de un paciente con factores de riesgo atendido en el area prehospitalaria al sufrir arresto cardiaco y presentar fibrilacion ventricular. El paciente fue atendido bajo estandares de Soporte Basico Vital y Soporte Cardiovascular Avanzado Vital, se aplico un Desfibrilador Automatizado Externo (DEA con resultado favorable y exito al recuperar al paciente de su condicion de compromiso de vida. Este es el primer reporte documentado con resultado favorable en el pais, en el area prehospitalaria y refuerza la conveniencia de adoptar politicas de Acceso Publico a la Desfibrilacion Temprana.This case report relates out-of-hospital care to a patient with risk factors treated in the out-of-hospital services after cardiac arrest and ventricular fibrillation. The patient was treated according to the standards of basic life support and advanced cardiovascular life support; by applying an automated external defibrillator (AED with favorable outcome and successful recovery of the patient from his risk of life condition. This is the first documented report with a favorable outcome in Peru, in out-of-hospital services and stresses the desirability of adopting policies for public access to early defibrillation.

  10. An MRI-Conditional External Cardiac Defibrillator for Resuscitation Within the MRI Scanner Bore

    Science.gov (United States)

    Schmidt, Ehud J.; Watkins, Ronald D.; Zviman, Menekhem M.; Guttman, Michael A.; Wang, Wei; Halperin, Henry A.

    2016-01-01

    Background Subjects undergoing cardiac arrest within an MRI scanner are currently removed from the bore and then from the MRI suite, prior to delivery of CPR and defibrillation, potentially increasing risk of mortality. This precludes many higher-risk (acute-ischemic, acute-stroke) patients from undergoing MRI imaging and MRI-guided intervention. An MRI-conditional cardiac defibrillator should enable scanning with defibrillation pads attached and the generator ON, enabling application of defibrillation within the MRI seconds after a cardiac event. An MRI-conditional external defibrillator may improve patient acceptance for MRI procedures. Methods and Results A commercial external defibrillator was rendered 1.5 Tesla MRI-conditional by addition of novel Radio-Frequency (RF) filters between the generator and commercial disposable surface-pads. The RF filters reduced emission into the MRI scanner, and prevented cable/surface-pad heating during imaging, while preserving all the defibrillator’s monitoring and delivery functions. Human volunteers were imaged using high Specific-Absorption-Rate sequences to validate MRI image quality (IQ) and lack of heating. Swine were electrically fibrillated (N=4) and thereafter defibrillated both outside and inside the MRI bore. MRI IQ was reduced by 0.8 or 1.6 dB, with the generator in monitoring mode and operating on battery or AC power, respectively. Commercial surface-pads did not create artifacts deeper than 6mm below the skin surface. RF heating was within FDA guidelines. Defibrillation was completely successful inside and outside the MRI bore. Conclusions A prototype MRI-conditional defibrillation system successfully defibrillated in the MRI without degrading image quality, or increasing the time needed for defibrillation. It can increase patient acceptance for MRI procedures. PMID:27729363

  11. [Learning to use semiautomatic external defibrillators through audiovisual materials for schoolchildren].

    Science.gov (United States)

    Jorge-Soto, Cristina; Abelairas-Gómez, Cristian; Barcala-Furelos, Roberto; Gregorio-García, Carolina; Prieto-Saborit, José Antonio; Rodríguez-Núñez, Antonio

    2016-01-01

    To assess the ability of schoolchildren to use a automated external defibrillator (AED) to provide an effective shock and their retention of the skill 1 month after a training exercise supported by audiovisual materials. Quasi-experimental controlled study in 205 initially untrained schoolchildren aged 6 to 16 years old. SAEDs were used to apply shocks to manikins. The students took a baseline test (T0) of skill, and were then randomized to an experimental or control group in the first phase (T1). The experimental group watched a training video, and both groups were then retested. The children were tested in simulations again 1 month later (T2). A total of 196 students completed all 3 phases. Ninety-six (95.0%) of the secondary school students and 54 (56.8%) of the primary schoolchildren were able to explain what a SAED is. Twenty of the secondary school students (19.8%) and 8 of the primary schoolchildren (8.4%) said they knew how to use one. At T0, 78 participants (39.8%) were able to simulate an effective shock. At T1, 36 controls (34.9%) and 56 experimental-group children (60.2%) achieved an effective shock (Paudiovisual instruction improves students' skill in managing a SAED and helps them retain what they learned for later use.

  12. Public Access Defibrillation

    DEFF Research Database (Denmark)

    Agerskov, Marianne; Nielsen, Anne Møller; Hansen, Carolina Malta

    2015-01-01

    BACKGROUND: In Copenhagen, a volunteer-based Automated External Defibrillator (AED) network provides a unique opportunity to assess AED use. We aimed to determine the proportion of Out-of-Hospital Cardiac Arrest (OHCA) where an AED was applied before arrival of the ambulance, and the proportion o...

  13. A Magnetic Resonance Imaging-Conditional External Cardiac Defibrillator for Resuscitation Within the Magnetic Resonance Imaging Scanner Bore.

    Science.gov (United States)

    Schmidt, Ehud J; Watkins, Ronald D; Zviman, Menekhem M; Guttman, Michael A; Wang, Wei; Halperin, Henry A

    2016-10-01

    Subjects undergoing cardiac arrest within a magnetic resonance imaging (MRI) scanner are currently removed from the bore and then from the MRI suite, before the delivery of cardiopulmonary resuscitation and defibrillation, potentially increasing the risk of mortality. This precludes many higher-risk (acute ischemic and acute stroke) patients from undergoing MRI and MRI-guided intervention. An MRI-conditional cardiac defibrillator should enable scanning with defibrillation pads attached and the generator ON, enabling application of defibrillation within the seconds of MRI after a cardiac event. An MRI-conditional external defibrillator may improve patient acceptance for MRI procedures. A commercial external defibrillator was rendered 1.5 Tesla MRI-conditional by the addition of novel radiofrequency filters between the generator and commercial disposable surface pads. The radiofrequency filters reduced emission into the MRI scanner and prevented cable/surface pad heating during imaging, while preserving all the defibrillator monitoring and delivery functions. Human volunteers were imaged using high specific absorption rate sequences to validate MRI image quality and lack of heating. Swine were electrically fibrillated (n=4) and thereafter defibrillated both outside and inside the MRI bore. MRI image quality was reduced by 0.8 or 1.6 dB, with the generator in monitoring mode and operating on battery or AC power, respectively. Commercial surface pads did not create artifacts deeper than 6 mm below the skin surface. Radiofrequency heating was within US Food and Drug Administration guidelines. Defibrillation was completely successful inside and outside the MRI bore. A prototype MRI-conditional defibrillation system successfully defibrillated in the MRI without degrading the image quality or increasing the time needed for defibrillation. It can increase patient acceptance for MRI procedures. © 2016 American Heart Association, Inc.

  14. Differences between out-of-hospital cardiac arrest in residential and public locations and implications for public-access defibrillation

    DEFF Research Database (Denmark)

    Folke, Fredrik; Gislason, Gunnar H; Lippert, Freddy

    2010-01-01

    The majority of out-of-hospital cardiac arrests (OHCAs) occur in residential locations, but knowledge about strategic placement of automated external defibrillators in residential areas is lacking. We examined whether residential OHCA areas suitable for placement of automated external defibrillat...... defibrillators could be identified on the basis of demographic characteristics and characterized individuals with OHCA in residential locations....

  15. Ressuscitação cardiopulmonar com a utilização do desfibrilador externo semi-automático: avaliação do processo ensino-aprendizagem Resucitador cardiopulmonar con utilización del disfibrilador externo semiautomático: evaluación del proceso enseñanza-aprendizaje Cardiopulmonary resuscitation with semi-automated external defibrillator: assessment of the teaching-learning process

    Directory of Open Access Journals (Sweden)

    Ana Maria Kazue Miyadahira

    2008-09-01

    that, for every minute delayed on defibrillating a heart arrest patient, survival chances decrease by 10%, and that the same chances of survival are 98% effective when it is employed within 30 seconds. While attending a heart arrest patient, it is crucial that the use of external semi-automated defibrillator (AED is included in the training. The purpose of the present study is to compare Psychomotor Ability and the Theoretical Knowledge of lay people on cardiopulmonary resuscitation (CPR using AED, before and after training. This sample was composed of 40 administrative workers of a public institution that were trained on CPR technique using EAD, as an experiment. The significantly higher scores in the assessment instrument items of Psychomotor Ability and Theoretical Knowledge, after training, indicates that the participants have presented improvements in their performances.

  16. TED-Time and life saving External Defibrillator for home-use.

    Science.gov (United States)

    Weiss, Teddy A; Rosenheck, Shimon; Gorni, Shraga; Katz, Ioni; Mendelbaum, Mendel; Gilon, Dan

    2014-06-01

    Sudden Cardiac Death--SCD --is a major unmet health problem that needs urgent and prompt solution. AICDs are very expensive, risky and indicated for a small group of patients, at the highest risk. AEDs--Automatic External Defibrillators--are designed for public places and although safe, cannot enter the home-market due to their cost and need for constant, high-cost maintenance. We developed TED, a low-cost AED that derives its energy off the mains, designed for home-use, to save SCD victims' lives. Copyright © 2014 Elsevier Inc. All rights reserved.

  17. Public Claims about Automatic External Defibrillators: An Online Consumer Opinions Study

    Directory of Open Access Journals (Sweden)

    Barnett Julie

    2011-05-01

    Full Text Available Abstract Background Patients are no longer passive recipients of health care, and increasingly engage in health communications outside of the traditional patient and health care professional relationship. As a result, patient opinions and health related judgements are now being informed by a wide range of social, media, and online information sources. Government initiatives recognise self-delivery of health care as a valuable means of responding to the anticipated increased global demand for health resources. Automated External Defibrillators (AEDs, designed for the treatment of Sudden Cardiac Arrest (SCA, have recently become available for 'over the counter' purchase with no need for a prescription. This paper explores the claims and argumentation of lay persons and health care practitioners and professionals relating to these, and how these may impact on the acceptance, adoption and use of these devices within the home context. Methods We carry out a thematic content analysis of a novel form of Internet-based data: online consumer opinions of AED devices posted on Amazon.com, the world's largest online retailer. A total of #83 online consumer reviews of home AEDs are analysed. The analysis is both inductive, identifying themes that emerged from the data, exploring the parameters of public debate relating to these devices, and also driven by theory, centring around the parameters that may impact upon the acceptance, adoption and use of these devices within the home as indicated by the Technology Acceptance Model (TAM. Results Five high-level themes around which arguments for and against the adoption of home AEDs are identified and considered in the context of TAM. These include opinions relating to device usability, usefulness, cost, emotional implications of device ownership, and individual patient risk status. Emotional implications associated with AED acceptance, adoption and use emerged as a notable factor that is not currently reflected

  18. Public claims about automatic external defibrillators: an online consumer opinions study.

    Science.gov (United States)

    Money, Arthur G; Barnett, Julie; Kuljis, Jasna

    2011-05-18

    Patients are no longer passive recipients of health care, and increasingly engage in health communications outside of the traditional patient and health care professional relationship. As a result, patient opinions and health related judgements are now being informed by a wide range of social, media, and online information sources. Government initiatives recognise self-delivery of health care as a valuable means of responding to the anticipated increased global demand for health resources. Automated External Defibrillators (AEDs), designed for the treatment of Sudden Cardiac Arrest (SCA), have recently become available for 'over the counter' purchase with no need for a prescription. This paper explores the claims and argumentation of lay persons and health care practitioners and professionals relating to these, and how these may impact on the acceptance, adoption and use of these devices within the home context. We carry out a thematic content analysis of a novel form of Internet-based data: online consumer opinions of AED devices posted on Amazon.com, the world's largest online retailer. A total of #83 online consumer reviews of home AEDs are analysed. The analysis is both inductive, identifying themes that emerged from the data, exploring the parameters of public debate relating to these devices, and also driven by theory, centring around the parameters that may impact upon the acceptance, adoption and use of these devices within the home as indicated by the Technology Acceptance Model (TAM). Five high-level themes around which arguments for and against the adoption of home AEDs are identified and considered in the context of TAM. These include opinions relating to device usability, usefulness, cost, emotional implications of device ownership, and individual patient risk status. Emotional implications associated with AED acceptance, adoption and use emerged as a notable factor that is not currently reflected within the existing TAM. The value, credibility and

  19. AED (Automated External Defibrillator) Programs: Questions and Answers

    Science.gov (United States)

    ... In Schools Training Kits RQI AHA Blended Learning & eLearning Guide AHA Instructors ECC Educational Conferences Programs CPR ... In Schools Training Kits RQI AHA Blended Learning & eLearning Guide AHA Instructors ECC Educational Conferences Programs CPR ...

  20. Automated External Defibrillators: Do You Need an AED?

    Science.gov (United States)

    ... places, including malls, office buildings, sports arenas and airplanes. However, many cardiac arrests occur at home, so ... Accessed Jan. 30, 2017. April 19, 2017 Original article: http://www.mayoclinic.org/diseases-conditions/heart-arrhythmia/ ...

  1. Externalized conductors and insulation failure in Biotronik defibrillator leads: History repeating or a false alarm?

    Science.gov (United States)

    De Maria, Elia; Borghi, Ambra; Bonetti, Lorenzo; Fontana, Pier Luigi; Cappelli, Stefano

    2017-02-16

    Conductor externalization and insulation failure are frequent complications with the recalled St. Jude Medical Riata implantable cardioverter-defibrillator (ICD) leads. Conductor externalization is a "unique" failure mechanism: Cables externalize through the insulation ("inside-out" abrasion) and appear outside the lead body. Recently, single reports described a similar failure also for Biotronik leads. Moreover, some studies reported a high rate of electrical dysfunction (not only insulation failure) with Biotronik Linox leads and a reduced survival rate in comparison with the competitors. In this paper we describe the case of a patient with a Biotronik Kentrox ICD lead presenting with signs of insulation failure and conductor externalization at fluoroscopy. Due to the high risk of extraction we decided to implant a new lead, abandoning the damaged one; lead reimplant was uneventful. Subsequently, we review currently available literature about Biotronik Kentrox and Linox ICD lead failure and in particular externalized conductors. Some single-center studies and a non-prospective registry reported a survival rate between 88% and 91% at 5 years for Linox leads, significantly worse than that of other manufacturers. However, the preliminary results of two ongoing multicenter, prospective registries (GALAXY and CELESTIAL) showed 96% survival rate at 5 years after implant, well within industry standards. Ongoing data collection is needed to confirm longer-term performance of this family of ICD leads.

  2. Defibrillation probability and impedance change between shocks during resuscitation from out-of-hospital cardiac arrest

    NARCIS (Netherlands)

    Walker, Robert G.; Koster, Rudolph W.; Sun, Charles; Moffat, George; Barger, Joseph; Dodson, Pamela P.; Chapman, Fred W.

    2009-01-01

    Objective: Technical data now gathered by automated external defibrillators (AEDs) allows closer evaluation of the behavior of defibrillation shocks administered during out-of-hospital cardiac arrest. We analyzed technical data from a large case series to evaluate the change in transthoracic

  3. Transmural recording of shock potential gradient fields, early postshock activations, and refibrillation episodes associated with external defibrillation of long-duration ventricular fibrillation in swine.

    Science.gov (United States)

    Allred, James D; Killingsworth, Cheryl R; Allison, J Scott; Dosdall, Derek J; Melnick, Sharon B; Smith, William M; Ideker, Raymond E; Walcott, Gregory P

    2008-11-01

    Knowledge of the shock potential gradient (nablaV) and postshock activation is limited to internal defibrillation of short-duration ventricular fibrillation (SDVF). The purpose of this study was to determine these variables after external defibrillation of long-duration VF (LDVF). In six pigs, 115-20 plunge needles with three to six electrodes each were inserted to record throughout both ventricles. After the chest was closed, the biphasic defibrillation threshold (DFT) was determined after 20 seconds of SDVF with external defibrillation pads. After 7 minutes of LDVF, defibrillation shocks that were less than or equal to the SDVF DFT strength were given. For DFT shocks (1632 +/- 429 V), the maximum minus minimum ventricular voltage (160 +/- 100 V) was 9.8% of the shock voltage. Maximum cardiac nablaV (28.7 +/- 17 V/cm) was 4.7 +/- 2.0 times the minimum nablaV (6.2 +/- 3.5 V/cm). Although LDVF did not increase the DFT in five of the six pigs, it significantly lengthened the time to earliest postshock activation following defibrillation (1.6 +/- 2.2 seconds for SDVF and 4.9 +/- 4.3 seconds for LDVF). After LDVF, 1.3 +/- 0.8 episodes of spontaneous refibrillation occurred per animal, but there was no refibrillation after SDVF. Compared with previous studies of internal defibrillation, during external defibrillation much less of the shock voltage appears across the heart and the shock field is much more even; however, the minimum nablaV is similar. Compared with external defibrillation of SDVF, the biphasic external DFT for LDVF is not increased; however, time to earliest postshock activation triples. Refibrillation is common after LDVF but not after SDVF in these normal hearts, indicating that LDVF by itself can cause refibrillation without requiring preexisting heart disease.

  4. General anaesthesia for insertion of an automated implantable cardioverter defibrillator in a child with Brugada and autism

    Directory of Open Access Journals (Sweden)

    Shwetal Goraksha

    2010-01-01

    Full Text Available A 14-year-old autistic boy presented with acute gastroenteritis and hypotension. The electrocardiogram showed a ventricular fibrillation rhythm - he went into cardiorespiratory arrest and was immediately resuscitated. On investigation, the electrocardiogram showed a partial right bundle branch block with a "coved" pattern of ST elevation in leads v 1 -v 3 . A provisional diagnosis of Brugada syndrome was made, for which an automated implantable cardioverter defibrillator (AICD implantation was advised. Although the automated implantable cardioverter defibrillator implantation is usually performed under sedation, because this was an autistic child, he needed general anaesthesia. We performed the procedure uneventfully under general anaesthesia and he was discharged after a short hospital stay.

  5. Training lay-people to use automatic external defibrillators: are all of their needs being met?

    Science.gov (United States)

    Harrison-Paul, Russell; Timmons, Stephen; van Schalkwyk, Wilna Dirkse

    2006-10-01

    We explored the experiences of lay people who have been trained to use automatic external defibrillators. The research questions were: (1) How can training courses help prepare people for dealing with real life situations? (2) Who is ultimately responsible for providing critical incident debriefing and how should this be organised? (3) What is the best process for providing feedback to those who have used an AED? Fifty-three semi-structured, qualitative interviews were conducted, some with those who had been trained and others with trainers. Locations included airports, railway stations, private companies and first responder schemes. Geographically, we covered Nottinghamshire, Lincolnshire, Yorkshire, Staffordshire, Essex and the West Midlands in the UK. Our analysis of the data indicates that most people believe scenarios based within their place of work were most useful in preparing for 'real life'. Many people had not received critical incident debriefing after using an AED. There were a variety of systems in place to provide support after an incident, many of which were informal. Training scenarios should be conducted outside the classroom. There should be more focus on critical incident debriefing during training and a clear identification of who should provide support after an incident. Other issues which were of interest included: (1) people's views on do not attempt resuscitation (DNAR); (2) perceived boundaries of responsibility when using an AED; (3) when is someone no longer 'qualified' to use an AED?

  6. Genoplivning med automatisk ekstern defibrillator på hospital

    DEFF Research Database (Denmark)

    Løfgren, Bo; Wahlgreen, Claus; Hoffmann, Anne Mette

    2009-01-01

    Early defibrillation is a determinant of survival in cardiac arrest. We report a Danish case of successful in-hospital resuscitation using an automated external defibrillator (AED). This case illustrates important aspects of implementation of in-hospital use of an AED, i.e. location of the AED......, education of the staff, systematic registration and data collection and technical aspects of AED use. If in-hospital AED implementation is carefully executed, its use may provide a safe and effective way of obtaining early defibrillation. Udgivelsesdato: 2009-Jan-26...

  7. Is Distance to the Nearest Registered Public Automated Defibrillator Associated with the Probability of Bystander Shock for Victims of Out-of-Hospital Cardiac Arrest?

    Science.gov (United States)

    Neves Briard, Joel; de Montigny, Luc; Ross, Dave; de Champlain, François; Segal, Eli

    2018-04-01

    Introduction Rapid access to defibrillation is a key element in the management of out-of-hospital cardiac arrests (OHCAs). Public automated external defibrillators (PAEDs) are becoming increasingly available, but little information exists regarding the relation between the proximity to the arrest and their usage in urban areas. This study is a retrospective, observational, cross-sectional analysis of non-traumatic OHCA during a 24-month period in the greater Montreal area (Quebec, Canada). Using logistic regression, bystander shock odds are described with regards to distance from the OHCA scene to the nearest PAED, adjusted for prehospital care arrival delay and time of day, and stratifying for type of location. Out of a total of 2,443 OHCA victims identified, 77 (3%) received bystander PAED shock, 622 (26%) occurred out-of-home, and 743 (30%) occurred during business hours. When controlling for time (business hours versus other hours) and minimum response delay for prehospital care arrival, a marginal negative association was found between bystander shock and distance to the nearest PAED in logged meters (aOR=0.80; CI, 0.64-0.99) for out-of-home cardiac arrests. No significant association was found between distance and bystander shock for at-home arrests. Out-of-home victims had significantly higher odds of receiving bystander shock up to 175 meters of distance to a PAED inclusively (aOR=2.52; CI, 1.07-5.89). For out-of-home cardiac arrests, proximity to a PAED was associated with bystander shock in the greater Montreal area. Strategies aiming to increase accessibility and use of these life-saving devices could further expand this advantage by assisting bystanders in rapidly locating nearby PAEDs. Neves Briard J , de Montigny L , Ross D , de Champlain F , Segal E . Is distance to the nearest registered public automated defibrillator associated with the probability of bystander shock for victims of out-of-hospital cardiac arrest? Prehosp Disaster Med. 2018;33(2):153-159.

  8. Bystander defibrillation for out-of-hospital cardiac arrest in Public vs Residential Locations

    DEFF Research Database (Denmark)

    Hansen, Steen Møller; Hansen, Carolina Malta; Folke, Fredrik

    2017-01-01

    Importance: Bystander-delivered defibrillation (hereinafter referred to as bystander defibrillation) of patients with out-of-hospital cardiac arrests (OHCAs) remains limited despite the widespread dissemination of automated external defibrillators (AEDs). Objective: To examine calendar changes...... in bystander defibrillation and subsequent survival according to a public or a residential location of the cardiac arrest after nationwide initiatives in Denmark to facilitate bystander-mediated resuscitative efforts, including bystander defibrillation. Design, Setting, and Participants: This nationwide study......, 2016. Exposures: Nationwide initiatives to facilitate bystander resuscitative efforts, including bystander defibrillation, consisted of resuscitation training of Danish citizens, dissemination of on-site AEDs, foundation of an AED registry linked to emergency medical dispatch centers, and dispatcher...

  9. Hvor er den automatiske eksterne defibrillator? Udvikling og uddannelse

    DEFF Research Database (Denmark)

    Løfgren, Bo; Grove, Erik; Krarup, Niels Henrik

    2009-01-01

    The International Liaison Committee on Resuscitation has presented a universal sign to indicate the presence of automated external defibrillators (AED). To disseminate the knowledge of this important signage, a review in a Danish context is presented. It is essential that the public in general...

  10. Programa de capacitação em ressuscitação cardiorrespiratória com uso do desfibrilador externo automático em uma universidade Programa de formación en reanimación cardiopulmonar con el uso del desfibrilador externo automático en una universidad Training program on cardiopulmonary resuscitation with the use of automated external defibrillator in a university

    Directory of Open Access Journals (Sweden)

    Ana Paula Boaventura

    2012-03-01

    .Early defibrillation in cardiopulmonary resuscitation (CPR receives increasing emphasis on its priority and rapidity. This is an experience report about the implementation of a training program in CPR using a defibrillator in a private university. The training program in basic CPR maneuvers was based on global guidelines, including a theorical course with practical demonstration of CPR maneuvers with the defibrillator, individual practical training and theoretical and practical assessments. About the performance of students in the practical assessment, the mean scores obtained by students in the first stage of the course was 26.4 points, while in the second stage the mean was 252.8 points; in the theoretical assessment the mean in the first stage was 3.06 points and in the second 9.0 points. The implementation of programs like this contribute to the effective acquisition of knowledge (theory and skill (pratice for the care of CPR victims.

  11. Barriers and facilitators to public access defibrillation in out-of-hospital cardiac arrest: a systematic review.

    Science.gov (United States)

    Smith, Christopher M; Lim Choi Keung, Sarah N; Khan, Mohammed O; Arvanitis, Theodoros N; Fothergill, Rachael; Hartley-Sharpe, Christopher; Wilson, Mark H; Perkins, Gavin D

    2017-10-01

    Public access defibrillation initiatives make automated external defibrillators available to the public. This facilitates earlier defibrillation of out-of-hospital cardiac arrest victims and could save many lives. It is currently only used for a minority of cases. The aim of this systematic review was to identify barriers and facilitators to public access defibrillation. A comprehensive literature review was undertaken defining formal search terms for a systematic review of the literature in March 2017. Studies were included if they considered reasons affecting the likelihood of public access defibrillation and presented original data. An electronic search strategy was devised searching MEDLINE and EMBASE, supplemented by bibliography and related-article searches. Given the low-quality and observational nature of the majority of articles, a narrative review was performed. Sixty-four articles were identified in the initial literature search. An additional four unique articles were identified from the electronic search strategies. The following themes were identified related to public access defibrillation: knowledge and awareness; willingness to use; acquisition and maintenance; availability and accessibility; training issues; registration and regulation; medicolegal issues; emergency medical services dispatch-assisted use of automated external defibrillators; automated external defibrillator-locator systems; demographic factors; other behavioural factors. In conclusion, several barriers and facilitators to public access defibrillation deployment were identified. However, the evidence is of very low quality and there is not enough information to inform changes in practice. This is an area in urgent need of further high-quality research if public access defibrillation is to be increased and more lives saved. PROSPERO registration number CRD42016035543. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2017. For permissions

  12. The challenges and possibilities of public access defibrillation.

    Science.gov (United States)

    Ringh, M; Hollenberg, J; Palsgaard-Moeller, T; Svensson, L; Rosenqvist, M; Lippert, F K; Wissenberg, M; Malta Hansen, C; Claesson, A; Viereck, S; Zijlstra, J A; Koster, R W; Herlitz, J; Blom, M T; Kramer-Johansen, J; Tan, H L; Beesems, S G; Hulleman, M; Olasveengen, T M; Folke, F

    2018-03-01

    Out-of-hospital cardiac arrest (OHCA) is a major health problem that affects approximately four hundred and thousand patients annually in the United States alone. It is a major challenge for the emergency medical system as decreased survival rates are directly proportional to the time delay from collapse to defibrillation. Historically, defibrillation has only been performed by physicians and in-hospital. With the development of automated external defibrillators (AEDs), rapid defibrillation by nonmedical professionals and subsequently by trained or untrained lay bystanders has become possible. Much hope has been put to the concept of Public Access Defibrillation with a massive dissemination of public available AEDs throughout most Western countries. Accordingly, current guidelines recommend that AEDs should be deployed in places with a high likelihood of OHCA. Despite these efforts, AED use is in most settings anecdotal with little effect on overall OHCA survival. The major reasons for low use of public AEDs are that most OHCAs take place outside high incidence sites of cardiac arrest and that most OHCAs take place in residential settings, currently defined as not suitable for Public Access Defibrillation. However, the use of new technology for identification and recruitment of lay bystanders and nearby AEDs to the scene of the cardiac arrest as well as new methods for strategic AED placement redefines and challenges the current concept and definitions of Public Access Defibrillation. Existing evidence of Public Access Defibrillation and knowledge gaps and future directions to improve outcomes for OHCA are discussed. In addition, a new definition of the different levels of Public Access Defibrillation is offered as well as new strategies for increasing AED use in the society. © 2018 The Association for the Publication of the Journal of Internal Medicine.

  13. Cardiopulmonary resuscitation and automatic external defibrillator training in schools: "is anyone learning how to save a life?".

    Science.gov (United States)

    Hart, Devin; Flores-Medrano, Oscar; Brooks, Steve; Buick, Jason E; Morrison, Laurie J

    2013-09-01

    Bystander resuscitation efforts, such as cardiopulmonary resuscitation (CPR) and use of an automatic external defibrillator (AED), save lives in cardiac arrest cases. School training in CPR and AED use may increase the currently low community rates of bystander resuscitation. The study objective was to determine the rates of CPR and AED training in Toronto secondary schools and to identify barriers to training and training techniques. This prospective study consisted of telephone interviews conducted with key school staff knowledgeable about CPR and AED teaching. An encrypted Web-based tool with prespecified variables and built-in logic was employed to standardize data collection. Of 268 schools contacted, 93% were available for interview and 83% consented to participate. Students and staff were trained in CPR in 51% and 80% of schools, respectively. Private schools had the lowest training rate (39%). Six percent of schools provided AED training to students and 47% provided AED training to staff. Forty-eight percent of schools had at least one AED installed, but 25% were unaware if their AED was registered with emergency services dispatch. Cost (17%), perceived need (11%), and school population size (10%) were common barriers to student training. Frequently employed training techniques were interactive (32%), didactic instruction (30%) and printed material (16%). CPR training rates for staff and students were moderate overall and lowest in private schools, whereas training rates in AED use were poor in all schools. Identified barriers to training include cost and student population size (perceived to be too small to be cost-effective or too large to be implemented). Future studies should assess the application of convenient and cost-effective teaching alternatives not presently in use.

  14. 78 FR 17890 - Effective Date of Requirement for Premarket Approval for Automated External Defibrillator System.

    Science.gov (United States)

    2013-03-25

    ... requiring this device to meet the statute's premarket approval requirements and the benefits to the public... requiring the device to have an approved PMA or a declared completed PDP and the benefit to the public from... consider exercising enforcement discretion for devices lawfully distributed before the requirement to have...

  15. Basic life support and automated external defibrillator skills among ambulance personnel

    DEFF Research Database (Denmark)

    Nielsen, Anne Møller; Isbye, Dan Lou; Lippert, Freddy Knudsen

    2012-01-01

    Ambulance personnel play an essential role in the 'Chain of Survival'. The prognosis after out-of-hospital cardiac arrest was dismal on a rural Danish island and in this study we assessed the cardiopulmonary resuscitation performance of ambulance personnel on that island.......Ambulance personnel play an essential role in the 'Chain of Survival'. The prognosis after out-of-hospital cardiac arrest was dismal on a rural Danish island and in this study we assessed the cardiopulmonary resuscitation performance of ambulance personnel on that island....

  16. Out-of-hospital cardiac arrest and placement of automated external defibrillators in the community

    DEFF Research Database (Denmark)

    Folke, Fredrik

    2010-01-01

    forskelle i patientkarakteristika afhængigt af hvor hjertestoppet forekommer (boligområde versus offentligt rum); 4) at estimere omkostnings-effektiviteten for PAD programmer i boligområder og det offentligt rum afhængigt af valgte AED placeringsstrategi. METODE OG RESULTATER I perioden 1994-2005 blev alle...

  17. Bystander Defibrillation for Out-of-Hospital Cardiac Arrest in Public vs Residential Locations.

    Science.gov (United States)

    Hansen, Steen Møller; Hansen, Carolina Malta; Folke, Fredrik; Rajan, Shahzleen; Kragholm, Kristian; Ejlskov, Linda; Gislason, Gunnar; Køber, Lars; Gerds, Thomas A; Hjortshøj, Søren; Lippert, Freddy; Torp-Pedersen, Christian; Wissenberg, Mads

    2017-05-01

    Bystander-delivered defibrillation (hereinafter referred to as bystander defibrillation) of patients with out-of-hospital cardiac arrests (OHCAs) remains limited despite the widespread dissemination of automated external defibrillators (AEDs). To examine calendar changes in bystander defibrillation and subsequent survival according to a public or a residential location of the cardiac arrest after nationwide initiatives in Denmark to facilitate bystander-mediated resuscitative efforts, including bystander defibrillation. This nationwide study identified 18 688 patients in Denmark with first-time OHCA from June 1, 2001, to December 31, 2012, using the Danish Cardiac Arrest Registry. Patients had a presumed cardiac cause of arrest that was not witnessed by emergency medical services personnel. Data were analyzed from April 1, 2015, to December 10, 2016. Nationwide initiatives to facilitate bystander resuscitative efforts, including bystander defibrillation, consisted of resuscitation training of Danish citizens, dissemination of on-site AEDs, foundation of an AED registry linked to emergency medical dispatch centers, and dispatcher-assisted guidance of bystander resuscitation efforts. The proportion of patients who received bystander defibrillation according to the location of the cardiac arrest and their subsequent 30-day survival. Of the 18 688 patients with OHCAs (67.8% men and 32.2% women; median [interquartile range] age, 72 [62-80] years), 4783 (25.6%) had a cardiac arrest in a public location and 13 905 (74.4%) in a residential location. The number of registered AEDs increased from 141 in 2007 to 7800 in 2012. The distribution of AED location was consistently skewed in favor of public locations. Bystander defibrillation increased in public locations from 3 of 245 (1.2%; 95% CI, 0.4%-3.5%) in 2001 to 78 of 510 (15.3%; 95% CI, 12.4%-18.7%) in 2012 (P bystander defibrillation increased in public locations from 8.3% (95% CI, 1.5%-35.4%) in 2001/2002 to 57

  18. Use and benefits of public access defibrillation in a nation-wide network

    DEFF Research Database (Denmark)

    Nielsen, Anne Møller; Folke, Fredrik; Lippert, Freddy Knudsen

    2013-01-01

    BACKGROUND: Automated External Defibrillators (AEDs) are known to increase survival after out-of-hospital cardiac arrest (OHCA). The aim of this study was to examine the use and benefit of public-access defibrillation (PAD) in a nation-wide network. We primarily sought to assess survival at 1 month...... to exercise (42% vs. 0%), and with improved 30-day survival (69% vs. 15%, p=0.001). Among those presenting with a shockable rhythm, 20 (65%) had Return of Spontaneous Circulation upon arrival of EMS and 8 (26%) were conscious, which emphasizes the diagnostic value of ECG downloads from AEDs. Survival could...

  19. PA/Lateral chest X-ray is equivalent to cine-fluoroscopy for the detection of conductor externalization in defibrillation leads.

    Science.gov (United States)

    Steinberg, Christian; Sarrazin, Jean-François; Philippon, François; Champagne, Jean; Molin, Franck; Nault, Isabelle; Blier, Louis; Bouchard, Marc-André; Arsenault, Jean; O'Hara, Gilles

    2015-01-01

    Riata™ and Riata ST defibrillation leads (St. Jude Medical, Sylmar, CA, USA) are susceptible to insulation defects with conductor externalization. Cine-fluoroscopy is considered to be the gold standard for the documentation of insulation defects, but similar detection rates have been reported for posterior-anterior (PA)/lateral chest x-ray (CXR) with zooming. Prospective single-center study to assess the diagnostic equivalence of a PA/lateral CXR with zooming for the detection of Riata insulation defects in a direct comparison to cine-fluoroscopy. Seventy-eight consecutive patients underwent 3-view cine-fluoroscopy and a PA/lateral CXR. All CXRs and cine-fluoroscopy images were reviewed by blinded electrophysiologists and staff radiologists. Forty-four of 78 patients had an abnormal cine-fluoroscopy (56%). The diagnostic correlation between PA/lateral CXR and cine-fluoroscopy was excellent (κ = 0.90; 95% confidence interval 0.80-1.00). PA/lateral CXR was equivalent to cine-fluoroscopy for the detection of conductor externalization showing a sensitivity of 97.7% and a specificity of 91.2%. The mean radiation effective dose of CXR was significantly lower compared to cine-fluoroscopy (0.09 millisievert [mSV] vs 0.85 ± 0.47 mSv; P cine-fluoroscopy for the detection of Riata insulation defects and should be considered as the preferred screening method. ©2014 Wiley Periodicals, Inc.

  20. Manuell arytmitolkning och defibrillering prehospitalt för att minska avbrott i bröstkompressioner

    OpenAIRE

    Mattsson, Andreas; Erling, Kristofer

    2014-01-01

    Abstract Background Todays guidelines for advanced CPR emphasize chest compressions with good quality and early defibrillation. Prehospital CPR performed by ambulance crew, an automated external defibrillator (AED) is used. The AED analyzes the heart rhythm and the performer is following the advice to chock the heart or not, given by the AED. During on-going CPR there are sequences when no chest compression is performed known as hands-off time. Hands-off time includes the time for the AED to ...

  1. Basic life support and external defibrillation competences after instruction and at 6 months comparing face-to-face and blended training. Randomised trial.

    Science.gov (United States)

    Castillo, Jordi; Gallart, Aberto; Rodríguez, Encarnación; Castillo, Jorge; Gomar, Carmen

    2018-06-01

    The objective of this study was to compare the immediate and 6-month efficacy of basic life support (BLS) and automatic external defibrillation (AED) training using standard or blended methods. First-year students of medicine and nursing (n = 129) were randomly assigned to a control group (face-to-face training based on the European Resuscitation Council [ERC] Guidelines) or to an experimental group that trained with a self-training video, a new website, a Moodle platform, an intelligent manikin, and 45 min of instructor presence. Both groups were homogeneous and were evaluated identically. Theoretical knowledge was evaluated using a multi-choice questionnaire (MCQ). Skill performance was evaluated by the instructor's rubric and on a high-fidelity Resusci Anne QCPR manikin. Immediately after the course, there were no statistically significant differences in knowledge between the two groups. The median score of practical evaluation assessed by the instructor was significantly better in the experimental group (8.15, SD 0.93 vs 7.7, SD 1.18; P = 0.02). No differences between groups were found when using a high-fidelity manikin to evaluate chest compressions and lung inflations. At six months, the scores in knowledge and skill performance were significantly lower compared to the evaluations at the end of the instruction, but they remained still higher compared to baseline. The experimental group had higher scores in practical skills evaluated by the instructor than the control group (7.44, SD 1.85 vs 6.10, SD 2.6; P = 0.01). The blended method provides the same or even higher levels of knowledge and skills than standard instruction both immediately after the course and six months later. Copyright © 2018 Elsevier Ltd. All rights reserved.

  2. Early Outcomes of Out-of-Hospital Cardiac Arrest after Early Defibrillation: a 24 Months Retrospective Analysis

    Directory of Open Access Journals (Sweden)

    Paolo Terranova

    2006-10-01

    Full Text Available Introduction: Cardiovascular disease remains the most common cause of death in the United States and most other Western nations. Among these deaths, sudden, out-of-hospital cardiac arrest claims approximately 1000 lives each day in the United States alone. Most of these cardiac arrests are due to ventricular fibrillation. Though highly reversible with the rapid application of a defibrillator, ventricular fibrillation is otherwise fatal within minutes, even when cardiopulmonary resuscitation is provided immediately. The overall survival rate in the United States is estimated to be less than 5 percent. Recent developments in automated-external-defibrillator technology have provided a means of increasing the rate of prompt defibrillation after out-of-hospital cardiac arrest. After minimal training, nonmedical personnel (e.g., flight attendants and casino workers are also able to use defibrillators in the workplace, with lifesaving effects. Nonetheless, such programs have involved designated personnel whose job description includes assisting persons who have had sudden cardiac arrest. Data are still lacking on the success of programs in which automated external defibrillators have been installed in public places to be used by persons who have no specific training or duty to act. Materials and Methods: All patients who had an out-of-hospital cardiac arrest between January 2003 and December 2004 and who received early defibrillation for ventricular fibrillation were included. We conducted a 24 months retrospective population-based analysis of the outcome in our population. Results: Over a 24 month period, 446 people had non–traumatic cardiac arrest, and in all of them it was observed to be ventricular fibrillation. In a very few cases, the defibrillator operators were good Samaritans, acting voluntarily. Eighty-nine patients (about 19% with ventricular fibrillation were successfully resuscitated, including eighteen who regained consciousness before

  3. Ranking Businesses and Municipal Locations by Spatiotemporal Cardiac Arrest Risk to Guide Public Defibrillator Placement.

    Science.gov (United States)

    Sun, Christopher L F; Brooks, Steven C; Morrison, Laurie J; Chan, Timothy C Y

    2017-03-21

    Efforts to guide automated external defibrillator placement for out-of-hospital cardiac arrest (OHCA) treatment have focused on identifying broadly defined location categories without considering hours of operation. Broad location categories may be composed of many businesses with varying accessibility. Identifying specific locations for automated external defibrillator deployment incorporating operating hours and time of OHCA occurrence may improve automated external defibrillator accessibility. We aim to identify specific businesses and municipal locations that maximize OHCA coverage on the basis of spatiotemporal assessment of OHCA risk in the immediate vicinity of franchise locations. This study was a retrospective population-based cohort study using data from the Toronto Regional RescuNET Epistry cardiac arrest database. We identified all nontraumatic public OHCAs occurring in Toronto, ON, Canada, from January 2007 through December 2015. We identified 41 unique businesses and municipal location types with ≥20 locations in Toronto from the YellowPages, Canadian Franchise Association, and the City of Toronto Open Data Portal. We obtained their geographic coordinates and hours of operation from Web sites, by phone, or in person. We determined the number of OHCAs that occurred within 100 m of each location when it was open (spatiotemporal coverage) for Toronto overall and downtown. The businesses and municipal locations were then ranked by spatiotemporal OHCA coverage. To evaluate temporal stability of the rankings, we calculated intraclass correlation of the annual coverage values. There were 2654 nontraumatic public OHCAs. Tim Hortons ranked first in Toronto, covering 286 OHCAs. Starbucks ranked first in downtown, covering 110 OHCAs. Coffee shops and bank machines from the 5 largest Canadian banks occupied 8 of the top 10 spots in both Toronto and downtown. The rankings exhibited high temporal stability with intraclass correlation values of 0.88 (95

  4. The number of prehospital defibrillation shocks and 1-month survival in patients with out-of-hospital cardiac arrest.

    Science.gov (United States)

    Hasegawa, Manabu; Abe, Takeru; Nagata, Takashi; Onozuka, Daisuke; Hagihara, Akihito

    2015-04-17

    The relationship between the number of pre-hospital defibrillation shocks and treatment outcome in patients with out-of-hospital cardiac arrest (OHCA) presenting with ventricular fibrillation (VF) is unknown currently. We examined the association between the number of pre-hospitalization defibrillation shocks and 1-month survival in OHCA patients. We conducted a prospective observational study using national registry data obtained from patients with OHCA between January 1, 2009 and December 31, 2012 in Japan. The study subjects were ≥ 18-110 years of age, had suffered from an OHCA before arrival of EMS personnel, had a witnessed collapse, had an initial rhythm that was shockable [VF/ventricular tachycardia (pulseless VT)], were not delivered a shock using a public automated external defibrillator (AED), received one or more shocks using a biphasic defibrillator by EMS personnel, and were transported to a medical institution between January 1, 2009 and December 31, 2012. There were 20,851 OHCA cases which met the inclusion criteria during the study period. Signal detection analysis was used to identify the cutoff point in the number of prehospital defibrillation shocks most closely related to one-month survival. Variables related to the number of defibrillations or one-month survival in OHCA were identified using multiple logistic regression analysis. A cutoff point in the number of pre-hospital defibrillation shocks most closely associated with 1-month OHCA survival was between two and three (χ(2) = 209.61, p < 0.0001). Among those patients who received two shocks or less, 34.48% survived for at least 1 month, compared with 24.75% of those who received three shocks or more. The number of defibrillations (odds ratio [OR] = 1.19, 95% CI: 1.03, 1.38), OHCA origin (OR = 2.81, 95% CI: 2.26, 3.49), use of ALS devices (OR = 0.68, 95% CI: 0.59, 0.79), use of epinephrine (OR = 0.33, 95% C: 0.28, 0.39), interval between first defibrillation and first ROSC (OR = 1.45, 95

  5. Return of spontaneous circulation and long-term survival according to feedback provided by automated external defibrillators

    DEFF Research Database (Denmark)

    Agerskov, M; Hansen, M B; Nielsen, A M

    2017-01-01

    levels. METHODS: We collected data on OHCA occurring between 2011 and 2014 in the Capital Region of Denmark where an AED was applied prior to ambulance arrival. Patient data were obtained from the Danish Cardiac Arrest Registry and medical records. AED data were retrieved from the Emergency Medical...... Dispatch Centre (EMDC) and information on feedback mechanisms, energy waveform and energy level was downloaded from the applied AEDs. RESULTS: A total of 196 OHCAs had an AED applied prior to ambulance arrival; 62 of these (32%) provided audio visual (AV) feedback while no feedback was provided in 134 (68...

  6. Comparison of instructor-led automated external defibrillation training and three alternative DVD-based training methods

    NARCIS (Netherlands)

    de Vries, Wiebe; Turner, Nigel M.; Monsieurs, Koenraad G.; Bierens, Joost J. L. M.; Koster, Rudolph W.

    2010-01-01

    Background: Self-directed BLS-training, using a personal training manikin with video has been shown to be as effective as instructor-led training. This has not previously been investigated for AED-training. Materials and methods: This prospective, randomized study with a non-inferiority design

  7. Lay Bystanders' Perspectives on What Facilitates Cardiopulmonary Resuscitation and Use of Automated External Defibrillators in Real Cardiac Arrests

    DEFF Research Database (Denmark)

    Malta Hansen, Carolina; Rosenkranz, Simone Mørk; Folke, Fredrik

    2017-01-01

    factors encourage lay bystanders to initiate CPR and AED use in a cohort of bystanders previously trained in CPR techniques who were present at an out-of-hospital cardiac arrest. METHODS AND RESULTS: One-hundred and twenty-eight semistructured qualitative interviews with CPR-trained lay bystanders...... obligation to act. CONCLUSIONS: Several factors other than previous hands-on CPR training facilitate lay bystander instigation of CPR and AED use. The recognition and modification of these factors may increase lay bystander CPR rates and patient survival following an out-of-hospital cardiac arrest....

  8. The Effects of Public Access Defibrillation on Survival After Out-of-Hospital Cardiac Arrest: A Systematic Review of Observational Studies.

    Science.gov (United States)

    Bækgaard, Josefine S; Viereck, Søren; Møller, Thea Palsgaard; Ersbøll, Annette Kjær; Lippert, Freddy; Folke, Fredrik

    2017-09-05

    Despite recent advances, the average survival after out-of-hospital cardiac arrest (OHCA) remains 50%. Accordingly, placement of automated external defibrillators in the community as part of a public access defibrillation program (PAD) is recommended by international guidelines. However, different strategies have been proposed on how exactly to increase and make use of publicly available automated external defibrillators. This systematic review aimed to evaluate the effect of PAD and the different PAD strategies on survival after OHCA. PubMed, Embase, and the Cochrane Library were systematically searched on August 31, 2015 for observational studies reporting survival to hospital discharge in OHCA patients where an automated external defibrillator had been used by nonemergency medical services. PAD was divided into 3 groups according to who applied the defibrillator: nondispatched lay first responders, professional first responders (firefighters/police) dispatched by the Emergency Medical Dispatch Center (EMDC), or lay first responders dispatched by the EMDC. A total of 41 studies were included; 18 reported PAD by nondispatched lay first responders, 20 reported PAD by EMDC-dispatched professional first responders (firefighters/police), and 3 reported both. We identified no qualified studies reporting survival after PAD by EMDC-dispatched lay first responders. The overall survival to hospital discharge after OHCA treated with PAD showed a median survival of 40.0% (range, 9.1-83.3). Defibrillation by nondispatched lay first responders was associated with the highest survival with a median survival of 53.0% (range, 26.0-72.0), whereas defibrillation by EMDC-dispatched professional first responders (firefighters/police) was associated with a median survival of 28.6% (range, 9.0-76.0). A meta-analysis of the different survival outcomes could not be performed because of the large heterogeneity of the included studies. This systematic review showed a median overall

  9. A first city-wide early defibrillation project in a German city: 5-year results of the Bochum against sudden cardiac arrest study

    Directory of Open Access Journals (Sweden)

    Hanefeld Christoph

    2010-06-01

    Full Text Available Abstract Background Immediate defibrillation is the decisive determinant of prognosis in patients suffering from cardiac/circulatory arrest caused by ventricular fibrillation (VF. Therefore, various national and international associations recommend that first responders use defibrillators as soon as possible and also recommend public access to early defibrillation programmes. Here we report the results of the first city-wide early defibrillation project in a large German urban area. Methods There were 155 automated external defibrillators (AEDs put into operation in the Bochum municipal area, and 6,294 people took part in cardiopulmonary resuscitation (CPR and AED training. Free, accessible AEDs were installed in places with large volumes of people. Additionally, emergency forces were progressively equipped with AEDs. Results Twelve AED administrations prior to the arrival of an emergency physician were recorded and analysed over a period of 5 years (08/2004-08/2009. Rhythm analysis via AED demonstrated VF in seven cases, non-malignant dysrhythmias in four cases and asystole in one case. Two of the seven patients with VF were successfully defibrillated and survived cardiac/circulatory arrest without any neurological sequelae. Eight of the 12 AED applications were performed by laymen. The mean time between switching the unit on and applying the electrodes to the patient was 39 seconds (SD +/-20 sec. On average, another 20 seconds elapsed before the AED recommendation of "shock delivery" was displayed, and a total of 96 seconds elapsed before shock administration (± 56 sec. Conclusion Consistent with other reports, our findings show that the organisation of a city-wide initiative by a project office combining public access and first-responder defibrillation programmes can be safe, feasible and successful. Our experiences confirm that strategic planning of AED placement is a prerequisite for successful, cost-effective resuscitation.

  10. Implantable defibrillator therapy: more than defibrillation...

    NARCIS (Netherlands)

    D.A.M.J. Theuns (Dominic)

    2005-01-01

    textabstractDuring the past 25 years, the implantable cardioverter-defibrillator (ICD) has evolved from the treatment of last resort to the gold standard for patients at high risk for life­threatening ventricular tachyarrhythmias. Patients at high risk include those who survived life-threatening

  11. Definition of successful defibrillation

    NARCIS (Netherlands)

    Koster, Rudolph W.; Walker, Robert G.; van Alem, Anouk P.

    2006-01-01

    OBJECTIVES: The definition of defibrillation shock "success" endorsed by the International Liaison Committee on Resuscitation since the publication of Guidelines 2000 for Cardiopulmonary Resuscitation and Emergency Cardiac Care has been removal of ventricular fibrillation at 5 secs after shock

  12. Knowledge and attitudes towards cardiopulmonary resuscitation and defibrillation amongst Asian primary health care physicians

    Directory of Open Access Journals (Sweden)

    Marcus Eh Ong

    2009-11-01

    Full Text Available Marcus Eh Ong1, Susan Yap1, Kim P Chan1, Papia Sultana2, Venkataraman Anantharaman11Department of Emergency Medicine, 2Department of Clinical Research, Singapore General Hospital, SingaporeObjective: To assess the knowledge and attitudes of local primary health care physicians in relation to cardiopulmonary resuscitation (CPR and defibrillation.Methods: We conducted a survey on general practitioners in Singapore by using a self-administered questionnaire that comprised 29 questions.Results: The response rate was 80%, with 60 of 75 physicians completing the questionnaire. The average age of the respondents was 52 years. Sixty percent of them reported that they knew how to operate an automated external defibrillator (AED, and 38% had attended AED training. Only 36% were willing to perform mouth-to-mouth ventilation during CPR, and 53% preferred chest compression-only resuscitation (CCR to standard CPR. We found those aged <50 years were more likely to be trained in basic cardiac life support (BCLS (P < 0.001 and advanced cardiac life support (P = 0.005 or to have ever attended to a patient with cardiac arrest (P = 0.007. Female physicians tended to agree that all clinics should have AEDs (P = 0.005 and support legislation to make AEDs compulsory in clinics (P < 0.001. We also found that a large proportion of physicians who were trained in BCLS (P = 0.006 were willing to perform mouth-to-mouth ventilation.Conclusion: Most local primary care physicians realize the importance of defibrillation, and the majority prefer CCR to standard CPR.Keywords: general practitioners, cardiac arrest, resuscitation, defibrillation, attitude, knowledge

  13. Will medical examination gloves protect rescuers from defibrillation voltages during hands-on defibrillation?

    Science.gov (United States)

    Sullivan, Joseph L; Chapman, Fred W

    2012-12-01

    Continuing compressions during a defibrillation shock has been proposed as a method of reducing pauses in cardiopulmonary resuscitation (CPR) but the safety of this procedure is unproven. The medical examination gloves worn by rescuers play an important role in protecting the rescuer yet the electrical characteristics of these gloves are unknown. This study examined the response of medical examination gloves to defibrillation voltages. Part 1 of this study measured voltage-current curves for a small sample (8) of gloves. Part 2 tested more gloves (460) to determine the voltage required to produce a specific amount of current flow. Gloves were tested at two current levels: 0.1 mA and 10 mA. Testing included four glove materials (chloroprene, latex, nitrile, and vinyl) in a single layer and double-gloved. All gloves tested in part 1 allowed little current to flow (gloves and 93 of 120 (77%) double gloves allowed at least 0.1 mA of current flow at voltages within the external defibrillation voltage range. Also, 6 of 80 (7.5%) single gloves and 5 of 80 (6.2%) double gloves allowed over 10 mA. Few of the gloves tested limited the current to levels proven to be safe. A lack of sensation during hands-on defibrillation does not guarantee that a safety margin exists. As such, we encourage rescuers to minimize rather than eliminate the pause in compressions for defibrillation. Copyright © 2012 Elsevier Ireland Ltd. All rights reserved.

  14. Automated Microscopy: Macro Language Controlling a Confocal Microscope and its External Illumination: Adaptation for Photosynthetic Organisms.

    Science.gov (United States)

    Steinbach, Gábor; Kaňa, Radek

    2016-04-01

    Photosynthesis research employs several biophysical methods, including the detection of fluorescence. Even though fluorescence is a key method to detect photosynthetic efficiency, it has not been applied/adapted to single-cell confocal microscopy measurements to examine photosynthetic microorganisms. Experiments with photosynthetic cells may require automation to perform a large number of measurements with different parameters, especially concerning light conditions. However, commercial microscopes support custom protocols (through Time Controller offered by Olympus or Experiment Designer offered by Zeiss) that are often unable to provide special set-ups and connection to external devices (e.g., for irradiation). Our new system combining an Arduino microcontroller with the Cell⊕Finder software was developed for controlling Olympus FV1000 and FV1200 confocal microscopes and the attached hardware modules. Our software/hardware solution offers (1) a text file-based macro language to control the imaging functions of the microscope; (2) programmable control of several external hardware devices (light sources, thermal controllers, actuators) during imaging via the Arduino microcontroller; (3) the Cell⊕Finder software with ergonomic user environment, a fast selection method for the biologically important cells and precise positioning feature that reduces unwanted bleaching of the cells by the scanning laser. Cell⊕Finder can be downloaded from http://www.alga.cz/cellfinder. The system was applied to study changes in fluorescence intensity in Synechocystis sp. PCC6803 cells under long-term illumination. Thus, we were able to describe the kinetics of phycobilisome decoupling. Microscopy data showed that phycobilisome decoupling appears slowly after long-term (>1 h) exposure to high light.

  15. Azygos Vein Lead Implantation For High Defibrillation Thresholds In Implantable Cardioverter Defibrillator Placement

    Directory of Open Access Journals (Sweden)

    Naga VA Kommuri

    2010-01-01

    Full Text Available Evaluation of defibrillation threshold is a standard of care during implantation of implantable cardioverter defibrillator. High defibrillation thresholds are often encountered and pose a challenge to electrophysiologists to improve the defibrillation threshold. We describe a case series where defibrillation thresholds were improved after implanting a defibrillation lead in the azygos vein.

  16. Saving lives with public access defibrillation: A deadly game of hide and seek.

    Science.gov (United States)

    Sidebottom, David B; Potter, Ryan; Newitt, Laura K; Hodgetts, Gillian A; Deakin, Charles D

    2018-07-01

    Early defibrillation is a critical link in the chain of survival. Public access defibrillation (PAD) programmes utilising automated external defibrillators (AEDs) aim to decrease the time-to-first-shock, and improve survival from out-of-hospital cardiac arrest. Effective use of PADs requires rapid location of the device, facilitated by adequate signage. We aimed to therefore assess the quality of signage for PADs in the community. From April 2017 to January 2018 we surveyed community PADs available for public use on the 'Save a Life' AED locator mobile application in and around Southampton, UK. Location and signage characteristics were collected, and the distance from the furthest sign to the AED was measured. Researchers evaluated 201 separate PADs. All devices visited were included in the final analysis. No signage at all was present for 135 (67.2%) devices. Only 15/201 (7.5%) AEDs had signage at a distance from AED itself. In only 5 of these cases (2.5%) was signage mounted more than 5.0 m from the AED. When signage was present, 46 used 2008 ILCOR signage and 15 used 2006 Resuscitation Council (UK) signage. Signage visibility was partially or severely obstructed at 27/66 (40.9%) sites. None of the 45 GP surgeries surveyed used exterior signage or an exterior 24/7 access box. Current signage of PADs is poor and limits the device effectiveness by impeding public awareness and location of AEDs. Recommendations should promote visible signage within the operational radius of each AED. Copyright © 2018 Elsevier B.V. All rights reserved.

  17. Underutilisation of public access defibrillation is related to retrieval distance and time-dependent availability.

    Science.gov (United States)

    Deakin, Charles D; Anfield, Steve; Hodgetts, Gillian A

    2018-05-14

    Public access defibrillation doubles the chances of neurologically intact survival following out-of-hospital cardiac arrest (OHCA). Although there are increasing numbers of defibrillators (automated external defibrillator (AEDs)) available in the community, they are used infrequently, despite often being available. We aimed to match OHCAs with known AED locations in order to understand AED availability, the effects of reduced AED availability at night and the operational radius at which they can be effectively retrieved. All emergency calls to South Central Ambulance Service from April 2014 to April 2016 were screened to identify cardiac arrests. Each was mapped to the nearest AED, according to the time of day. Mapping software was used to calculate the actual walking distance for a bystander between each OHCA and respective AED, when travelling at a brisk walking speed (4 mph). 4012 cardiac arrests were identified and mapped to one of 2076 AEDs. All AEDs were available during daytime hours, but only 713 at night (34.3%). 5.91% of cardiac arrests were within a retrieval (walking) radius of 100 m during the day, falling to 1.59% out-of-hours. Distances to rural AEDs were greater than in urban areas (P<0.0001). An AED could potentially have been retrieved prior to actual ambulance arrival in 25.3% cases. Existing AEDs are underused; 36.4% of OHCAs are located within 500 m of an AED. Although more AEDs will improve availability, greater use can be made of existing AEDs, particularly by ensuring they are all available on a 24/7 basis. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018. All rights reserved. No commercial use is permitted unless otherwise expressly granted.

  18. Surviving out-of-hospital cardiac arrest: just a matter of defibrillators?

    Science.gov (United States)

    Zorzi, Alessandro; Gasparetto, Nicola; Stella, Federica; Bortoluzzi, Andrea; Cacciavillani, Luisa; Basso, Cristina

    2014-08-01

    Out-of-hospital sudden cardiac arrest (OHCA) is a leading cause of death all over the world. Although the outcome of OHCA resulting from 'nonshockable' rhythms (asystole and pulseless electrical activity) is poor regardless of resuscitation efforts, 'shockable' rhythms such as ventricular tachycardia or fibrillation may carry a good prognosis if early defibrillation is performed. At present, simplified cardiopulmonary resuscitation techniques (hands-only cardiopulmonary resuscitation) and automated external defibrillators (AEDs) offer lay people the possibility to provide lifesaving treatment to OHCA victims in the critical minutes before the arrival of the emergency medical system. Programs aimed at increasing provision of cardiopulmonary resuscitation and use of AEDs by lay people have been set up in different countries, including Italy, and have contributed to improve survival rates. However, success of these programs critically depends on appropriate planning and design, and on cultural predisposition of witnesses to undertake immediate measures of resuscitation in the case of OHCA. Placement of a large number of AEDs may carry high costs and little benefits if it is uncoordinated and not preceded by educational campaigns to spread widely the 'culture of resuscitation' in the population.

  19. 75 FR 70015 - External Defibrillators; Public Workshop

    Science.gov (United States)

    2010-11-16

    ... different environments (hospital, community, home)? 4. What features of next generation devices can be... present written material or to make oral presentations at the public workshop, please call the contact...

  20. Part 3: Adult Basic Life Support and Automated External Defibrillation 2015 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations

    NARCIS (Netherlands)

    Travers, Andrew H.; Perkins, Gavin D.; Berg, Robert A.; Castren, Maaret; Considine, Julie; Escalante, Raffo; Gazmuri, Raul J.; Koster, Rudolph W.; Lim, Swee Han; Nation, Kevin J.; Olasveengen, Theresa M.; Sakamoto, Tetsuya; Sayre, Michael R.; Sierra, Alfredo; Smyth, Michael A.; Stanton, David; Vaillancourt, Christian; Bierens, Joost J. L. M.; Bourdon, Emmanuelle; Brugger, Hermann; Buick, Jason E.; Charette, Manya L.; Chung, Sung Phil; Couper, Keith; Daya, Mohamud R.; Drennan, Ian R.; Gräsner, Jan-Thorsten; Idris, Ahamed H.; Lerner, E. Brooke; Lockhat, Husein; Løfgren, Bo; McQueen, Carl; Monsieurs, Koenraad G.; Mpotos, Nicolas; Orkin, Aaron M.; Quan, Linda; Raffay, Violetta; Reynolds, Joshua C.; Ristagno, Giuseppe; Scapigliati, Andrea; Vadeboncoeur, Tyler F.; Wenzel, Volker; Yeung, Joyce

    2015-01-01

    This review comprises the most extensive literature search and evidence evaluation to date on the most important international BLS interventions, diagnostics, and prognostic factors for cardiac arrest victims. It reemphasizes that the critical lifesaving steps of BLS are (1) prevention, (2)

  1. Temporal Trends in Coverage of Historical Cardiac Arrests Using a Volunteer-Based Network of Automated External Defibrillators Accessible to Laypersons and Emergency Dispatch Centers

    DEFF Research Database (Denmark)

    Hansen, Carolina Malta; Lippert, Freddy Knudsen; Wissenberg, Mads

    2014-01-01

    public cardiac arrest coverage in high- and low-risk areas. METHODS AND RESULTS: All public cardiac arrests (1994-2011) and all registered AEDs (2007-2011) in Copenhagen, Denmark, were identified and geocoded. AED coverage of cardiac arrests was defined as historical arrests ≤100 m from an AED. High...

  2. What externally presented information do VRUs require when interacting with fully Automated Road Transport Systems in shared space?

    Science.gov (United States)

    Merat, Natasha; Louw, Tyron; Madigan, Ruth; Wilbrink, Marc; Schieben, Anna

    2018-03-31

    As the desire for deploying automated ("driverless") vehicles increases, there is a need to understand how they might communicate with other road users in a mixed traffic, urban, setting. In the absence of an active and responsible human controller in the driving seat, who might currently communicate with other road users in uncertain/conflicting situations, in the future, understanding a driverless car's behaviour and intentions will need to be relayed via easily comprehensible, intuitive and universally intelligible means, perhaps presented externally via new vehicle interfaces. This paper reports on the results of a questionnaire-based study, delivered to 664 participants, recruited during live demonstrations of an Automated Road Transport Systems (ARTS; SAE Level 4), in three European cities. The questionnaire sought the views of pedestrians and cyclists, focussing on whether respondents felt safe interacting with ARTS in shared space, and also what externally presented travel behaviour information from the ARTS was important to them. Results showed that most pedestrians felt safer when the ARTS were travelling in designated lanes, rather than in shared space, and the majority believed they had priority over the ARTS, in the absence of such infrastructure. Regardless of lane demarcations, all respondents highlighted the importance of receiving some communication information about the behaviour of the ARTS, with acknowledgement of their detection by the vehicle being the most important message. There were no clear patterns across the respondents, regarding preference of modality for these external messages, with cultural and infrastructural differences thought to govern responses. Generally, however, conventional signals (lights and beeps) were preferred to text-based messages and spoken words. The results suggest that until these driverless vehicles are able to provide universally comprehensible externally presented information or messages during interaction

  3. Automated Microscopy: Macro Language Controlling a Confocal Microscope and its External Illumination: Adaptation for Photosynthetic Organisms

    Czech Academy of Sciences Publication Activity Database

    Steinbach, Gabor; Kaňa, Radek

    2016-01-01

    Roč. 22, č. 2 (2016), s. 258-263 ISSN 1431-9276 R&D Projects: GA ČR GAP501/12/0304; GA MŠk EE2.3.30.0059; GA MŠk ED2.1.00/03.0110; GA MŠk(CZ) LO1416 Institutional support: RVO:61388971 Keywords : automated microscopy * remote controlled microscopy * confocal microscopy Subject RIV: BH - Optics, Masers, Lasers Impact factor: 1.891, year: 2016

  4. A fully automated non-external marker 4D-CT sorting algorithm using a serial cine scanning protocol.

    Science.gov (United States)

    Carnes, Greg; Gaede, Stewart; Yu, Edward; Van Dyk, Jake; Battista, Jerry; Lee, Ting-Yim

    2009-04-07

    Current 4D-CT methods require external marker data to retrospectively sort image data and generate CT volumes. In this work we develop an automated 4D-CT sorting algorithm that performs without the aid of data collected from an external respiratory surrogate. The sorting algorithm requires an overlapping cine scan protocol. The overlapping protocol provides a spatial link between couch positions. Beginning with a starting scan position, images from the adjacent scan position (which spatial match the starting scan position) are selected by maximizing the normalized cross correlation (NCC) of the images at the overlapping slice position. The process was continued by 'daisy chaining' all couch positions using the selected images until an entire 3D volume was produced. The algorithm produced 16 phase volumes to complete a 4D-CT dataset. Additional 4D-CT datasets were also produced using external marker amplitude and phase angle sorting methods. The image quality of the volumes produced by the different methods was quantified by calculating the mean difference of the sorted overlapping slices from adjacent couch positions. The NCC sorted images showed a significant decrease in the mean difference (p < 0.01) for the five patients.

  5. External Validity of Electronic Sniffers for Automated Recognition of Acute Respiratory Distress Syndrome.

    Science.gov (United States)

    McKown, Andrew C; Brown, Ryan M; Ware, Lorraine B; Wanderer, Jonathan P

    2017-01-01

    Automated electronic sniffers may be useful for early detection of acute respiratory distress syndrome (ARDS) for institution of treatment or clinical trial screening. In a prospective cohort of 2929 critically ill patients, we retrospectively applied published sniffer algorithms for automated detection of acute lung injury to assess their utility in diagnosis of ARDS in the first 4 ICU days. Radiographic full-text reports were searched for "edema" OR ("bilateral" AND "infiltrate") and a more detailed algorithm for descriptions consistent with ARDS. Patients were flagged as possible ARDS if a radiograph met search criteria and had a PaO 2 /FiO 2 or SpO 2 /FiO 2 of 300 or 315, respectively. Test characteristics of the electronic sniffers and clinical suspicion of ARDS were compared to a gold standard of 2-physician adjudicated ARDS. Thirty percent of 2841 patients included in the analysis had gold standard diagnosis of ARDS. The simpler algorithm had sensitivity for ARDS of 78.9%, specificity of 52%, positive predictive value (PPV) of 41%, and negative predictive value (NPV) of 85.3% over the 4-day study period. The more detailed algorithm had sensitivity of 88.2%, specificity of 55.4%, PPV of 45.6%, and NPV of 91.7%. Both algorithms were more sensitive but less specific than clinician suspicion, which had sensitivity of 40.7%, specificity of 94.8%, PPV of 78.2%, and NPV of 77.7%. Published electronic sniffer algorithms for ARDS may be useful automated screening tools for ARDS and improve on clinical recognition, but they are limited to screening rather than diagnosis because their specificity is poor.

  6. Automated calculation of myocardial external efficiency from a single 11C-acetate PET/CT scan

    DEFF Research Database (Denmark)

    Harms, Hans; Tolbod, Lars Poulsen; Hansson, Nils Henrik

    of this study was to develop and validate an automated method of calculating MEE from a single dynamic 11C-acetate PETscan. Methods: 21 subjects underwent a dynamic 27 min 11C-acetate PETscan on a Siemens Biograph TruePoint 64 PET/CTscanner. Using cluster analysis, the LV-aortic time-activity curve (TACLV......). Conclusion: Myocardial efficiencycanbe derived directly andautomatically froma single dynamic 11C-acetate PET scan. This eliminates the need for a separate CMR scan and eliminates any potential errors due to different loading conditions between CMR and PETscans.......Background: Dynamic PETwith 11C-acetate can be used to assess myocardial oxygen use which in turn is usedto calculate myocardial external efficiency (MEE), anearly marker of heart failure. MEE is defined as the ratio of total work (TW) and total energy use (TE). Calculation of TW and TE requires...

  7. Desfibrilador automático implantado en pacientes con miocardiopatía hipertrófica: criterios de selección, evolución y predictores de terapia apropiada Automatic defibrillator implanted in patients with hypertrophic myocardiopathy: selection criteria, evolution and appropriate therapy predictors

    Directory of Open Access Journals (Sweden)

    Guillermo Mora

    Full Text Available Introducción y objetivos: la miocardiopatía hipertrófica es una enfermedad de origen genético con prevalencia de 1% al 2%. La mitad de los pacientes fallecen por muerte súbita cardiaca, la mayoría por arritmias ventriculares. Todavía no está claro a qué pacientes se les debe implantar un desfibrilador automático. El objetivo de este trabajo es describir una serie de pacientes con implante, los criterios empleados y los resultados obtenidos, así como analizar los predictores de terapia apropiada por el desfibrilador. Métodos: se incluyeron 20 pacientes que recibieron un desfibrilador de tercera generación. En todos se realizó estudio electrofisiológico y seguimiento prospectivo con registro de eventos. En 18 (90% se hizo estudio genético. Resultados: el 55% eran hombres con edad promedio de 40 [11-78] años. Seis (30% recibieron implantante por prevención secundaria y 14 (70% por prevención primaria; los últimos por presentar varios factores de riesgo. Se indujo una arritmia sostenida en 15 (75% y en 3 (15% taquicardia ventricular monomórfica sostenida. A 22 meses de seguimiento 4 (20% sufrieron terapia apropiada y 2 (10% fallecieron. La taquicardia ventricular monomórfica clínica (p=0,03 y la inducida (pIntroduction and objectives: hypertrophic myocardiopathy is a genetic entity with 1% to 2% prevalence. Half patients die of sudden cardiac death, most due to ventricular arrhythmias. There is still no clarity with regard to the patients to whom an automatic defibrillator has to be implanted. The objective of this work is to describe a series of patients with implant, the criteria used and the results obtained, as well as to analyze the predictors of appropriate therapy with the defibrillator. Methods: 20 patients that received a third generation defibrillator were included. Electrophysiological study and prospective follow-up with register of events was performed in all. Genetic study was done in 18 (90%. Results: 55% were

  8. Access to automatic defibrillation at airports on an example of Warsaw Chopin Airport

    Directory of Open Access Journals (Sweden)

    Witold Pawłowski

    2017-08-01

    Full Text Available Sudden cardiac arrest and cessation of blood circulation is the most common cause of death of people around the world. Immediate notification of emergency services and cardiopulmonary resuscitation combined with an automatic external defibrillator (AED increases the chances of survivors. Warsaw Chopin Airport is the only public place in Poland and the third one in Europe where a complex and integrated life saving system has been implemented in the ICC. The paper presents an analysis of the access to automatic defibrillation at airports at the Warsaw Chopin Airport

  9. Internal defibrillation: where we have been and where we should be going?

    Science.gov (United States)

    Lévy, Samuel

    2005-08-01

    Internal cardioversion has been developed as an alternative technique for patients who are resistant to external DC cardioversion of atrial fibrillation (AF) and was found to be associated with higher success rates. It used initially high energies (200-300 J) delivered between an intracardiac catheter and a backplate. Subsequent studies have shown that it is possible to terminate with energies of 1 to 6 Joules, paroxysmal or induced AF in 90 percent of patients and persistent AF in 75 percent of patients, using biphasic shocks delivered between a right atrium-coronary sinus vectors. Consequently, internal atrial defibrillation can be performed under sedation only without the need for general anesthesia. Recently developed external defibrillators, capable of delivering biphasic shocks, have increased the success rates of external cardioversion and reduced the need for internal cardioversion. However, internal defibrillation is still useful in overweight or obese patients, in patients with chronic obstructive pulmonary disease or asthma who are more difficult to defibrillate, and in patients with implanted devices which may be injured by high energy shocks. Low energy internal defibrillation has also proven to be safe and this has prompted the development of implantable devices for terminating AF. The first device used was the Metrix system, a stand-alone atrial defibrillator (without ventricular defibrillation) which was found to be safe and effective in selected groups of patients. Unfortunately, this device is no longer being marketed. Only double chamber defibrillators with pacing capabilities are presently available: the Medtronic GEM III AT, an updated version of the Jewel AF and the Guidant PRIZM AVT. These devices can be patient-activated or programmed to deliver automatically ounce atrial tachyarrhythmias are detected, therapies including pacing or/and shocks. Attempts to define the group of patients who might benefit from these devices are described but the

  10. A second defibrillator chest patch electrode will increase implantation rates for nonthoracotomy defibrillators.

    Science.gov (United States)

    Solomon, A J; Swartz, J F; Rodak, D J; Moore, H J; Hannan, R L; Tracy, C M; Fletcher, R D

    1996-09-01

    Nonthoracotomy defibrillator systems can be implanted with a lower morbidity and mortality, compared to epicardial systems. However, implantation may be unsuccessful in up to 15% of patients, using a monophasic waveform. It was the purpose of this study to prospectively examine the efficacy of a second chest patch electrode in a nonthoracotomy defibrillator system. Fourteen patients (mean age 62 +/- 11 years, ejection fraction = 0.29 +/- 0.12) with elevated defibrillation thresholds, defined as > or = 24 J, were studied. The initial lead system consisted of a right ventricular electrode (cathode), a left innominate vein, and subscapular chest patch electrode (anodes). If the initial defibrillation threshold was > or = 24 J, a second chest patch electrode was added. This was placed subcutaneously in the anterior chest (8 cases), or submuscularly in the subscapular space (6 cases). This resulted in a decrease in the system impedance at the defibrillation threshold, from 72.3 +/- 13.3 omega to 52.2 +/- 8.6 omega. Additionally, the defibrillation threshold decreased from > or = 24 J, with a single patch, to 16.6 +/- 2.8 J with two patches. These changes were associated with successful implantation of a nonthoracotomy defibrillator system in all cases. In conclusion, the addition of a second chest patch electrode (using a subscapular approach) will result in lower defibrillation thresholds in patients with high defibrillation thresholds, and will subsequently increase implantation rates for nonthoracotomy defibrillators.

  11. Optimal Implantable Cardioverter Defibrillator Programming.

    Science.gov (United States)

    Shah, Bindi K

    Optimal programming of implantable cardioverter defibrillators (ICDs) is essential to appropriately treat ventricular tachyarrhythmias and to avoid unnecessary and inappropriate shocks. There have been a series of large clinical trials evaluating tailored programming of ICDs. We reviewed the clinical trials evaluating ICD therapies and detection, and the consensus statement on ICD programming. In doing so, we found that prolonged ICD detection times, higher rate cutoffs, and antitachycardia pacing (ATP) programming decreases inappropriate and painful therapies in a primary prevention population. The use of supraventricular tachyarrhythmia discriminators can also decrease inappropriate shocks. Tailored ICD programming using the knowledge gained from recent ICD trials can decrease inappropriate and unnecessary ICD therapies and decrease mortality.

  12. The role of conductivity discontinuities in design of cardiac defibrillation

    Science.gov (United States)

    Lim, Hyunkyung; Cun, Wenjing; Wang, Yue; Gray, Richard A.; Glimm, James

    2018-01-01

    Fibrillation is an erratic electrical state of the heart, of rapid twitching rather than organized contractions. Ventricular fibrillation is fatal if not treated promptly. The standard treatment, defibrillation, is a strong electrical shock to reinitialize the electrical dynamics and allow a normal heart beat. Both the normal and the fibrillatory electrical dynamics of the heart are organized into moving wave fronts of changing electrical signals, especially in the transmembrane voltage, which is the potential difference between the cardiac cellular interior and the intracellular region of the heart. In a normal heart beat, the wave front motion is from bottom to top and is accompanied by the release of Ca ions to induce contractions and pump the blood. In a fibrillatory state, these wave fronts are organized into rotating scroll waves, with a centerline known as a filament. Treatment requires altering the electrical state of the heart through an externally applied electrical shock, in a manner that precludes the existence of the filaments and scroll waves. Detailed mechanisms for the success of this treatment are partially understood, and involve local shock-induced changes in the transmembrane potential, known as virtual electrode alterations. These transmembrane alterations are located at boundaries of the cardiac tissue, including blood vessels and the heart chamber wall, where discontinuities in electrical conductivity occur. The primary focus of this paper is the defibrillation shock and the subsequent electrical phenomena it induces. Six partially overlapping causal factors for defibrillation success are identified from the literature. We present evidence in favor of five of these and against one of them. A major conclusion is that a dynamically growing wave front starting at the heart surface appears to play a primary role during defibrillation by critically reducing the volume available to sustain the dynamic motion of scroll waves; in contrast, virtual

  13. Interaction of defibrillation waveform with the time to defibrillation or the number of defibrillation attempts on survival from out-of-hospital cardiac arrest.

    Science.gov (United States)

    Hagihara, Akihito; Onozuka, Daisuke; Ono, Junko; Nagata, Takashi; Hasegawa, Manabu

    2018-01-01

    Early biphasic defibrillation is effective in out-of-hospital cardiac arrest (OHCA) cases. In the resuscitation of patients with OHCA, it is not clear how the defibrillation waveform interacts with the time to defibrillation to influence patient survival. The second, and any subsequent, shocks need to be administered by an on-line physician in Japan. Thus, we investigated the interaction between the defibrillation waveform and time to or the number of defibrillation on resuscitation outcomes. This prospective observational study used data for all OHCAs that occurred between 2005 and 2014 in Japan. To investigate the interaction effect between the defibrillation waveform and the time to defibrillation or the number of defibrillations on the return to spontaneous circulation (ROSC), 1-month survival, and cerebral performance category (CPC) (1, 2), we assessed the modifying effects of the defibrillation waveform and the time to or the number of defibrillation on additive scale (i.e., the relative excessive risk due to interaction, RERI) and multiplicative scale (i.e., ratio of odds ratios (ORs)). In total, 71,566 cases met the inclusion criteria. For the measure of interaction between the defibrillation waveform and the time to defibrillation, ratio of ORs for ROSC was 0.84 (0.75-0.94), implying that the effect of time to first defibrillation on ROSC was negatively modified by defibrillation waveform. For the interaction between the defibrillation waveform and the number of defibrillations, RERI and ratio of ORs for CPC (1, 2) was -0.25 (-0.47 to -0.06) and 0.79 (0.67-0.93), respectively. It is implied that the effect of number of defibrillation on CPC (1, 2) was negatively modified by defibrillation waveform. An increased number of defibrillations was associated with a decreased ROSC in the case of biphasic and monophasic defibrillation, while an increased number of defibrillations was related to an increased 1-month survival rate and CPC (1, 2) only in the case of

  14. Towards Low Energy Atrial Defibrillation

    Directory of Open Access Journals (Sweden)

    Philip Walsh

    2015-09-01

    Full Text Available A wireless powered implantable atrial defibrillator consisting of a battery driven hand-held radio frequency (RF power transmitter (ex vivo and a passive (battery free implantable power receiver (in vivo that enables measurement of the intracardiac impedance (ICI during internal atrial defibrillation is reported. The architecture is designed to operate in two modes: Cardiac sense mode (power-up, measure the impedance of the cardiac substrate and communicate data to the ex vivo power transmitter and cardiac shock mode (delivery of a synchronised very low tilt rectilinear electrical shock waveform. An initial prototype was implemented and tested. In low-power (sense mode, >5 W was delivered across a 2.5 cm air-skin gap to facilitate measurement of the impedance of the cardiac substrate. In high-power (shock mode, >180 W (delivered as a 12 ms monophasic very-low-tilt-rectilinear (M-VLTR or as a 12 ms biphasic very-low-tilt-rectilinear (B-VLTR chronosymmetric (6ms/6ms amplitude asymmetric (negative phase at 50% magnitude shock was reliably and repeatedly delivered across the same interface; with >47% DC-to-DC (direct current to direct current power transfer efficiency at a switching frequency of 185 kHz achieved. In an initial trial of the RF architecture developed, 30 patients with AF were randomised to therapy with an RF generated M-VLTR or B-VLTR shock using a step-up voltage protocol (50–300 V. Mean energy for successful cardioversion was 8.51 J ± 3.16 J. Subsequent analysis revealed that all patients who cardioverted exhibited a significant decrease in ICI between the first and third shocks (5.00 Ω (SD(σ = 1.62 Ω, p < 0.01 while spectral analysis across frequency also revealed a significant variation in the impedance-amplitude-spectrum-area (IAMSA within the same patient group (|∆(IAMSAS1-IAMSAS3[1 Hz − 20 kHz] = 20.82 Ω-Hz (SD(σ = 10.77 Ω-Hz, p < 0.01; both trends being absent in all patients that failed to cardiovert

  15. Ranking businesses and municipal locations by spatiotemporal cardiac arrest risk to guide public defibrillator placement

    Science.gov (United States)

    Sun, Christopher L. F.; Brooks, Steven C.; Morrison, Laurie J.; Chan, Timothy C.Y.

    2017-01-01

    Background Efforts to guide automated external defibrillator (AED) placement for out-of-hospital cardiac arrest (OHCA) treatment have focused on identifying broadly defined location categories without considering hours of operation. Broad location categories may be composed of many businesses with varying accessibility. Identifying specific locations for AED deployment incorporating operating hours and time of OHCA occurrence may improve AED accessibility. We aim to identify specific businesses and municipal locations that maximize OHCA coverage based on spatiotemporal assessment of OHCA risk in the immediate vicinity of franchise locations. Methods This study was a retrospective population-based cohort study using data from the Toronto Regional RescuNET Epistry cardiac arrest database. We identified all non-traumatic public OHCAs occurring in Toronto, Canada from Jan. 2007–Dec. 2015. We identified 41 unique businesses and municipal location types with 20 or more locations in Toronto from the YellowPages, Canadian Franchise Association, and the City of Toronto Open Data Portal. We obtained their geographic coordinates and hours of operation from websites, phone, or in-person. We determined the number of OHCAs that occurred within 100 m of each location when it was open (spatiotemporal coverage) for Toronto overall and downtown. The businesses and municipal locations were then ranked by spatiotemporal OHCA coverage. To evaluate temporal stability of the rankings, we calculated intra-class correlation (ICC) of the annual coverage values. Results There were 2,654 non-traumatic public OHCAs. Tim Hortons ranked first in Toronto covering 286 OHCAs. Starbucks ranked first in downtown covering 110 OHCAs. Coffee shops and bank machines from the five largest Canadian banks occupied eight of the top 10 spots in both Toronto and downtown. The rankings exhibited high temporal stability with ICC values of 0.88 (95% CI, 0.83–0.93) in Toronto and 0.79 (95% CI, 0.71–0.86) in

  16. Intra-operative defibrillation testing and clinical shock efficacy in patients with implantable cardioverter-defibrillators

    DEFF Research Database (Denmark)

    Bänsch, Dietmar; Bonnemeier, Hendrik; Brandt, Johan

    2015-01-01

    AIMS: This trial was designed to test the hypothesis that shock efficacy during follow-up is not impaired in patients implanted without defibrillation (DF) testing during first implantable cardioverter-defibrillator (ICD) implantation. METHODS AND RESULTS: Between February 2011 and July 2013, 107...

  17. Automated Gain Control and Internal Calibration With External Ion Accumulation Capillary liquid chromatography-electrospray ionization-fourier transform ion cyclotron resonance.

    Energy Technology Data Exchange (ETDEWEB)

    Belov, Mikhail E.(VISITORS); Zhang, Rui (BATTELLE (PACIFIC NW LAB)); Strittmatter, Eric F.(BATTELLE (PACIFIC NW LAB)); Prior, David C.(BATTELLE (PACIFIC NW LAB)); Tang, Keqi (BATTELLE (PACIFIC NW LAB)); Smith, Richard D.(BATTELLE (PACIFIC NW LAB))

    2003-08-15

    When combined with capillary LC separations, Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry (ESI-FTICR MS) has increasingly been applied for advanced characterization of proteolytic digests. Incorporation of external (to the ICR cell) ion accumulation multipoles with FTICR for ion pre selection and accumulation has enhanced the dynamic range, sensitivity and duty cycle of measurements. However, the highly variable ion production rate from an LC separation can result in overfilling of the external trap, resulting in m/z discrimination and fragmentation of peptide ions. An excessive space charge trapped in the ICR cell causes significant shifts in the detected ion cyclotron frequencies, reducing the achievable mass measurement accuracy (MMA) for protein identification. To eliminate m/z discrimination in the external ion trap, further increase the duty cycle and improve MMA, we developed a capability for data-dependent adjustment of ion accumulation times in the course of an LC separation, referred to as Automated Gain Control (AGC), in combination with low kinetic energy gated ion trapping and internal calibration using a dual-channel electrodynamic ion funnel. The system was initially evaluated in the analysis of a 0.5 mg/mL tryptic digest of bovine serum albumin. The implementation of LC/ESI/AGC/FTICR with internal calibration gave rise to a {approx} 10-fold increase in the number of identified tryptic peptides within mass measurement accuracy of 2 ppm as compared to that detected during the conventional LC/FTICR run with a fixed ion accumulation time and external calibration.

  18. Application of an automated natural language processing (NLP) workflow to enable federated search of external biomedical content in drug discovery and development.

    Science.gov (United States)

    McEntire, Robin; Szalkowski, Debbie; Butler, James; Kuo, Michelle S; Chang, Meiping; Chang, Man; Freeman, Darren; McQuay, Sarah; Patel, Jagruti; McGlashen, Michael; Cornell, Wendy D; Xu, Jinghai James

    2016-05-01

    External content sources such as MEDLINE(®), National Institutes of Health (NIH) grants and conference websites provide access to the latest breaking biomedical information, which can inform pharmaceutical and biotechnology company pipeline decisions. The value of the sites for industry, however, is limited by the use of the public internet, the limited synonyms, the rarity of batch searching capability and the disconnected nature of the sites. Fortunately, many sites now offer their content for download and we have developed an automated internal workflow that uses text mining and tailored ontologies for programmatic search and knowledge extraction. We believe such an efficient and secure approach provides a competitive advantage to companies needing access to the latest information for a range of use cases and complements manually curated commercial sources. Copyright © 2016. Published by Elsevier Ltd.

  19. Vernakalant selectively prolongs atrial refractoriness with no effect on ventricular refractoriness or defibrillation threshold in pigs.

    Science.gov (United States)

    Bechard, Jeff; Gibson, John Ken; Killingsworth, Cheryl R; Wheeler, Jeffery J; Schneidkraut, Marlowe J; Huang, Jian; Ideker, Raymond E; McAfee, Donald A

    2011-03-01

    Vernakalant is a novel antiarrhythmic agent that has demonstrated clinical efficacy for the treatment of atrial fibrillation. Vernakalant blocks, to various degrees, cardiac sodium and potassium channels with a pattern that suggests atrial selectivity. We hypothesized, therefore, that vernakalant would affect atrial more than ventricular effective refractory period (ERP) and have little or no effect on ventricular defibrillation threshold (DFT). Atrial and ventricular ERP and ventricular DFT were determined before and after treatment with vernakalant or vehicle in 23 anesthetized male mixed-breed pigs. Vernakalant was infused at a rate designed to achieve stable plasma levels similar to those in human clinical trials. Atrial and ventricular ERP were determined by endocardial extrastimuli delivered to the right atria or right ventricle. Defibrillation was achieved using external biphasic shocks delivered through adhesive defibrillation patches placed on the thorax after 10 seconds of electrically induced ventricular fibrillation. The DFT was estimated using the Dixon "up-and-down" method. Vernakalant significantly increased atrial ERP compared with vehicle controls (34 ± 8 versus 9 ± 7 msec, respectively) without significantly affecting ventricular ERP or DFT. This is consistent with atrial selective actions and supports the conclusion that vernakalant does not alter the efficacy of electrical defibrillation.

  20. Rise and Shock: Optimal Defibrillator Placement in a High-rise Building.

    Science.gov (United States)

    Chan, Timothy C Y

    2017-01-01

    Out-of-hospital cardiac arrests (OHCA) in high-rise buildings experience lower survival and longer delays until paramedic arrival. Use of publicly accessible automated external defibrillators (AED) can improve survival, but "vertical" placement has not been studied. We aim to determine whether elevator-based or lobby-based AED placement results in shorter vertical distance travelled ("response distance") to OHCAs in a high-rise building. We developed a model of a single-elevator, n-floor high-rise building. We calculated and compared the average distance from AED to floor of arrest for the two AED locations. We modeled OHCA occurrences using floor-specific Poisson processes, the risk of OHCA on the ground floor (λ 1 ) and the risk on any above-ground floor (λ). The elevator was modeled with an override function enabling direct travel to the target floor. The elevator location upon override was modeled as a discrete uniform random variable. Calculations used the laws of probability. Elevator-based AED placement had shorter average response distance if the number of floors (n) in the building exceeded three quarters of the ratio of ground-floor OHCA risk to above-ground floor risk (λ 1 /λ) plus one half (n ≥ 3λ 1 /4λ + 0.5). Otherwise, a lobby-based AED had shorter average response distance. If OHCA risk on each floor was equal, an elevator-based AED had shorter average response distance. Elevator-based AEDs travel less vertical distance to OHCAs in tall buildings or those with uniform vertical risk, while lobby-based AEDs travel less vertical distance in buildings with substantial lobby, underground, and nearby street-level traffic and OHCA risk.

  1. Validation of defibrillator lead performance registry data

    DEFF Research Database (Denmark)

    Kristensen, Anders Elgaard; Larsen, Jacob Moesgaard; Nielsen, Jens Cosedis

    2017-01-01

    all reported surgical interventions due to defibrillator lead events in the Danish Pacemaker and ICD Register (DPIR) from 2000 to 2013. Medical records of all patients (n = 753) were examined blinded for 5 predefined intervention types and 18 reasons for lead intervention. The overall level...

  2. Do clinical examination gloves provide adequate electrical insulation for safe hands-on defibrillation? I: Resistive properties of nitrile gloves.

    Science.gov (United States)

    Deakin, Charles D; Lee-Shrewsbury, Victoria; Hogg, Kitwani; Petley, Graham W

    2013-07-01

    Uninterrupted chest compressions are a key factor in determining resuscitation success. Interruptions to chest compression are often associated with defibrillation, particularly the need to stand clear from the patient during defibrillation. It has been suggested that clinical examination gloves may provide adequate electrical resistance to enable safe hands-on defibrillation in order to minimise interruptions. We therefore examined whether commonly used nitrile clinical examination gloves provide adequate resistance to current flow to enable safe hands-on defibrillation. Clinical examination gloves (Kimberly Clark KC300 Sterling nitrile) worn by members of hospital cardiac arrest teams were collected immediately following termination of resuscitation. To determine the level of protection afforded by visually intact gloves, electrical resistance across the glove was measured by applying a DC voltage across the glove and measuring subsequent resistance. Forty new unused gloves (control) were compared with 28 clinical (non-CPR) gloves and 128 clinical (CPR) gloves. One glove in each group had a visible tear and was excluded from analysis. Control gloves had a minimum resistance of 120 kΩ (median 190 kΩ) compared with 60 kΩ in clinical gloves (both CPR (median 140 kΩ) and non-CPR groups (median 160 kΩ)). Nitrile clinical examination gloves do not provide adequate electrical insulation for the rescuer to safely undertake 'hands-on' defibrillation and when exposed to the physical forces of external chest compression, even greater resistive degradation occurs. Further work is required to identify gloves suitable for safe use for 'hands-on' defibrillation. Copyright © 2013 Elsevier Ireland Ltd. All rights reserved.

  3. Concerns about the implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; van Domburg, Ron T; Theuns, Dominic A M J

    2005-01-01

    Patients with an implantable cardioverter defibrillator (ICD) are at increased risk of anxiety disorders. In turn, anxiety has been identified as a precipitant of ventricular arrhythmias. Anxiety may in part be attributed to concerns about the ICD firing, but the relationship between ICD concerns......, psychological morbidity, and shocks has not been systematically investigated. We examined the relative importance of experienced shocks versus subjective concerns about the ICD as determinants of anxiety and depressive symptoms in ICD patients....

  4. Fast Electrocardiogram Amplifier Recovery after Defibrillation Shock

    Directory of Open Access Journals (Sweden)

    Ivan Dotsinsky

    2005-04-01

    Full Text Available A procedure for fast ECG amplifier recovery after defibrillation shocks was developed and simulated in the MATLAB environment. Exponentially decaying post-shock voltages have been recorded. Signals from the AHA database are taken and mixed with the recorded exponential disturbances. The algorithm applies moving averaging (comb filter on the compound input signal, thereby obtaining the samples of the disturbance. They are currently subtracted from the input signal. The results obtained show that its recovery is practically instantaneous.

  5. Single-Coil Defibrillator Leads Yield Satisfactory Defibrillation Safety Margin in Hypertrophic Cardiomyopathy.

    Science.gov (United States)

    Okamura, Hideo; Friedman, Paul A; Inoue, Yuko; Noda, Takashi; Aiba, Takeshi; Yasuda, Satoshi; Ogawa, Hisao; Kamakura, Shiro; Kusano, Kengo; Espinosa, Raul E

    2016-09-23

    Single-coil defibrillator leads have gained favor because of their potential ease of extraction. However, a high defibrillation threshold remains a concern in patients with hypertrophic cardiomyopathy (HCM), and in many cases, dual-coil leads have been used for this patient group. There is little data on using single-coil leads for HCM patients. We evaluated 20 patients with HCM who received an implantable cardioverter-defibrillator (ICD) on the left side in combination with a dual-coil lead. Two sets of defibrillation tests were performed in each patient, one with the superior vena cava (SVC) coil "on" and one with the SVC coil "off". ICDs were programmed to deliver 25 joules (J) for the first attempt followed by maximum energy (35 J or 40 J). Shock impedance and shock pulse width at 25 J in each setting as well as the results of the shock were analyzed. All 25-J shocks in both settings successfully terminated ventricular fibrillation. However, shock impedance and pulse width increased substantially with the SVC coil programmed "off" compared with "on" (66.4±6.1 ohm and 14.0±1.3 ms "off" vs. 41.9±5.0 ohm and 9.3±0.8 ms "on", Psatisfactory safety margin for 35-J devices. Single-coil leads appear appropriate for left-sided implantation in this patient group. (Circ J 2016; 80: 2199-2203).

  6. Impact of basic life-support training on the attitudes of health-care workers toward cardiopulmonary resuscitation and defibrillation.

    Science.gov (United States)

    Abolfotouh, Mostafa A; Alnasser, Manal A; Berhanu, Alamin N; Al-Turaif, Deema A; Alfayez, Abdulrhman I

    2017-09-22

    Cardiopulmonary resuscitation (CPR) increases the probability of survival of a person with cardiac arrest. Repeating training helps staff retain knowledge in CPR and in use of automated external defibrillators (AEDs). Retention of knowledge and skills during and after training in CPR is difficult and requires systematic training with appropriate methodology. The aim of this study was to determine the effect of basic life-support (BLS) training on the attitudes of health-care providers toward initiating CPR and on use of AEDs, and to investigate the factors that influence these attitudes. A quasi-experimental study was conducted in two groups: health-care providers who had just attended a BLS-AED course (post-BLS group, n = 321), and those who had not (pre-BLS group, n = 421). All participants had previously received BLS training. Both groups were given a validated questionnaire to evaluate the status of life-support education and certification, attitudes toward use of CPR and AED and concerns regarding use of CPR and AED. Multiple linear regression analyses were applied to identify significant predictors of the attitude and concern scores. Overall positive attitudes were seen in 53.4% of pre-BLS respondents and 64.8% of post-BLS respondents (χ 2  = 9.66, p = 0.002). Positive attitude was significantly predicted by the recent completion of BLS training (β = 5.15, p attitudes toward CPR performance and the use of AEDs. Training that addressed the concerns of health-care workers could further improve these attitudes.

  7. Electromagnetic interference from welding and motors on implantable cardioverter-defibrillators as tested in the electrically hostile work site.

    Science.gov (United States)

    Fetter, J G; Benditt, D G; Stanton, M S

    1996-08-01

    This study was designed to determine the susceptibility of an implanted cardioverter-defibrillator to electromagnetic interference in an electrically hostile work site environment, with the ultimate goal of allowing the patient to return to work. Normal operation of an implanted cardioverter-defibrillator depends on reliable sensing of the heart's electrical activity. Consequently, there is concern that external electromagnetic interference from external sources in the work place, especially welding equipment or motor-generator systems, may be sensed and produce inappropriate shocks or abnormal reed switch operation, temporarily suspending detection of ventricular tachycardia or ventricular fibrillation. The effects of electromagnetic interference on the operation of one type of implantable cardioverter-defibrillator (Medtronic models 7217 and 7219) was measured by using internal event counter monitoring in 10 patients operating arc welders at up to 900 A or working near 200-hp motors and 1 patient close to a locomotive starter drawing up to 400 A. The electromagnetic interference produced two sources of potential interference on the sensing circuit or reed switch operation, respectively: 1) electrical fields with measured frequencies up to 50 MHz produced by the high currents during welding electrode activation, and 2) magnetic fields produced by the current in the welding electrode and cable. The defibrillator sensitivity was programmed to the highest (most sensitive) value: 0.15 mV (model 7219) or 0.3 mV (model 7217). The ventricular tachycardia and ventricular fibrillation therapies were temporarily turned off but the detection circuits left on. None of the implanted defibrillators tested were affected by oversensing of the electric field as verified by telemetry from the detection circuits. The magnetic field from 225-A welding current produced a flux density of 1.2 G; this density was not adequate to close the reed switch, which requires approximately 10 G

  8. Extended charge banking model of dual path shocks for implantable cardioverter defibrillators.

    Science.gov (United States)

    Dosdall, Derek J; Sweeney, James D

    2008-08-01

    Single path defibrillation shock methods have been improved through the use of the Charge Banking Model of defibrillation, which predicts the response of the heart to shocks as a simple resistor-capacitor (RC) circuit. While dual path defibrillation configurations have significantly reduced defibrillation thresholds, improvements to dual path defibrillation techniques have been limited to experimental observations without a practical model to aid in improving dual path defibrillation techniques. The Charge Banking Model has been extended into a new Extended Charge Banking Model of defibrillation that represents small sections of the heart as separate RC circuits, uses a weighting factor based on published defibrillation shock field gradient measures, and implements a critical mass criteria to predict the relative efficacy of single and dual path defibrillation shocks. The new model reproduced the results from several published experimental protocols that demonstrated the relative efficacy of dual path defibrillation shocks. The model predicts that time between phases or pulses of dual path defibrillation shock configurations should be minimized to maximize shock efficacy. Through this approach the Extended Charge Banking Model predictions may be used to improve dual path and multi-pulse defibrillation techniques, which have been shown experimentally to lower defibrillation thresholds substantially. The new model may be a useful tool to help in further improving dual path and multiple pulse defibrillation techniques by predicting optimal pulse durations and shock timing parameters.

  9. Higher defibrillation threshold in methamphetamine cardiomyopathy patients with implantable cardioverter-defibrillator

    Directory of Open Access Journals (Sweden)

    Rakesh Malhotra

    2017-11-01

    Full Text Available Introduction: Identification of patients with an increased risk of high defibrillation thresholds (DFTs is important in planning implantable cardioverter-defibrillator (ICD procedures. Clinical observations have suggested that patients with methamphetamine cardiomyopathy (MACMP have significantly elevated defibrillation thresholds. We hypothesized that MACMP patients would have higher DFT thresholds than controls and would require procedural changes during ICD implantation to accommodate higher thresholds. Methods: We identified consecutive patients with MACMP undergoing ICD implantation at the academic center from 2003 to 2007. We then compared DFTs against age-and sex-matched controls. Results: The MACMP (n = 10 group showed significantly increased DFT thresholds (23.7 ± 6.7 J compared with age and sex-matched controls (14.5 ± 4.6 J, p < 0.005. Additionally, patients with MACMP had evidence of more severe congestive heart failure, with increased B-type natrieutic protein (BNP levels (1173 ± 784 vs 260 ± 349, p = 0.02 and decreased left ventricular ejection fraction (LVEF (17.8 ± 9.4 vs 35.9 ± 15.2, p = 0.02. MACMP patients required high output devices than controls (50% versus 0%, p = 0.03. Differences between groups remained significant despite adjusting for LVEF. Conclusions: Planning for ICD implantation should take into consideration a history of methamphetamine abuse, mandating DFT testing and empiric consideration of high output devices for such patients. Keywords: Methamphetamine cardiomyopathy, Implantable cardioverter-defibrillatior, Defibrillation threshold testing, B-type natriuretic peptide, Ejection fraction

  10. Psychological intervention following implantation of an implantable defibrillator

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; van den Broek, Krista C; Sears, Samuel F

    2007-01-01

    The medical benefits of the implantable cardioverter defibrillator (ICD) are unequivocal, but a subgroup of patients experiences emotional difficulties following implantation. For this subgroup, some form of psychological intervention may be warranted. This review provides an overview of current ...

  11. Implantable cardioverter defibrillator implantation in children in The Netherlands

    NARCIS (Netherlands)

    Ten Harkel, ADJ; Blom, NA; Reimer, AG; Tukkie, R; Sreeram, N; Bink-Boelkens, MTE

    To evaluate the indications, underlying cardiac disorders, efficacy and complications involved with implantable cardioverter-defibrillators (ICDs) in paediatric patients in The Netherlands, the records of all patients aged 18 years or younger who underwent ICD placement were reviewed

  12. Implantable cardioverter defibrillator implantation in children in The Netherlands

    NARCIS (Netherlands)

    ten Harkel, A. Derk Jan; Blom, Nico A.; Reimer, Annette G.; Tukkie, Raymond; Sreeram, Narayanswami; Bink-Boelkens, Margreet T. E.

    2005-01-01

    To evaluate the indications, underlying cardiac disorders, efficacy and complications involved with implantable cardioverter-defibrillators (ICDs) in paediatric patients in The Netherlands, the records of all patients aged 18 years or younger who underwent ICD placement were reviewed

  13. A patch in the pectoral position lowers defibrillation threshold.

    Science.gov (United States)

    Karasik, P; Solomon, A; Verdino, R; Moore, H; Rodak, D; Hannan, R; Fletcher, R

    1997-06-01

    Implantable pacemaker cardioverter defibrillators are now available with biphasic waveforms, which have been shown to markedly improve defibrillation thresholds (DFTs). However, in a number of patients the DFT remains high. Also, DFT may increase after implantation, especially if antiarrhythmic drugs are added. We report on the use of a subcutaneous patch in the pectoral position in 15 patients receiving a transvenous defibrillator as a method of easily reducing the DFT. A 660-mm2 patch electrode was placed beneath the generator in a pocket created on the pectoral fascia. The energy required for defibrillation was lowered by 56% on average, and the system impedance was lowered by a mean of 25%. This maneuver allowed all patients to undergo a successful implant with adequate safety margin.

  14. Elektrokirurgi hos patienter med pacemaker og implanterbar kardioverter-defibrillator

    DEFF Research Database (Denmark)

    Lønnberg, Ann Sophie Claire; Philberts, Berit Thornvig; Bonde, Christian

    2017-01-01

    Electrosurgery is a very useful tool and one of the most commonly used techniques. However, the technique can interfere with pacemakers and implantable cardioverter defibrillators. This article provides practical recommendations for the use of electrosurgery in these patients.......Electrosurgery is a very useful tool and one of the most commonly used techniques. However, the technique can interfere with pacemakers and implantable cardioverter defibrillators. This article provides practical recommendations for the use of electrosurgery in these patients....

  15. Gender differences in anxiety and concerns about the cardioverter defibrillator

    DEFF Research Database (Denmark)

    Spindler, Helle; Johansen, Jens B; Andersen, Kirsten Krogh

    2009-01-01

    Little is known about gender differences in the response to implantable cardioverter defibrillator (ICD) therapy. We compared female and male ICD patients on anxiety, depression, health-related quality of life (HRQL), ICD concerns, and ICD acceptance.......Little is known about gender differences in the response to implantable cardioverter defibrillator (ICD) therapy. We compared female and male ICD patients on anxiety, depression, health-related quality of life (HRQL), ICD concerns, and ICD acceptance....

  16. Impact of carvedilol and metoprolol on inappropriate implantable cardioverter-defibrillator therapy

    DEFF Research Database (Denmark)

    Ruwald, Martin H; Abu-Zeitone, Abeer; Jons, Christian

    2013-01-01

    The goal of this study was to evaluate the effects of carvedilol and metoprolol on the endpoint of inappropriate implantable cardioverter-defibrillator therapy in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation With Cardiac Resynchronization Therapy) study....

  17. Computational cardiology: the bidomain based modified Hill model incorporating viscous effects for cardiac defibrillation

    Science.gov (United States)

    Cansız, Barış; Dal, Hüsnü; Kaliske, Michael

    2017-10-01

    Working mechanisms of the cardiac defibrillation are still in debate due to the limited experimental facilities and one-third of patients even do not respond to cardiac resynchronization therapy. With an aim to develop a milestone towards reaching the unrevealed mechanisms of the defibrillation phenomenon, we propose a bidomain based finite element formulation of cardiac electromechanics by taking into account the viscous effects that are disregarded by many researchers. To do so, the material is deemed as an electro-visco-active material and described by the modified Hill model (Cansız et al. in Comput Methods Appl Mech Eng 315:434-466, 2017). On the numerical side, we utilize a staggered solution method, where the elliptic and parabolic part of the bidomain equations and the mechanical field are solved sequentially. The comparative simulations designate that the viscoelastic and elastic formulations lead to remarkably different outcomes upon an externally applied electric field to the myocardial tissue. Besides, the achieved framework requires significantly less computational time and memory compared to monolithic schemes without loss of stability for the presented examples.

  18. Pacemakers and implantable cardioverter defibrillators - general and anesthetic considerations

    Directory of Open Access Journals (Sweden)

    Amy G. Rapsang

    2014-06-01

    Full Text Available A pacemaking system consists of an impulse generator and lead or leads to carry the electrical impulse to the patient's heart. Pacemaker and implantable cardioverter defibrillator codes were made to describe the type of pacemaker or implantable cardioverter defibrillator implanted. Indications for pacing and implantable cardioverter defibrillator implantation were given by the American College of Cardiologists. Certain pacemakers have magnet-operated reed switches incorporated; however, magnet application can have serious adverse effects; hence, devices should be considered programmable unless known otherwise. When a device patient undergoes any procedure (with or without anesthesia, special precautions have to be observed including a focused history/physical examination, interrogation of pacemaker before and after the procedure, emergency drugs/temporary pacing and defibrillation, reprogramming of pacemaker and disabling certain pacemaker functions if required, monitoring of electrolyte and metabolic disturbance and avoiding certain drugs and equipments that can interfere with pacemaker function. If unanticipated device interactions are found, consider discontinuation of the procedure until the source of interference can be eliminated or managed and all corrective measures should be taken to ensure proper pacemaker function should be done. Post procedure, the cardiac rate and rhythm should be monitored continuously and emergency drugs and equipments should be kept ready and consultation with a cardiologist or a pacemaker-implantable cardioverter defibrillator service may be necessary.

  19. Regional Externalities

    NARCIS (Netherlands)

    Heijman, W.J.M.

    2007-01-01

    The book offers practical and theoretical insights in regional externalities. Regional externalities are a specific subset of externalities that can be defined as externalities where space plays a dominant role. This class of externalities can be divided into three categories: (1) externalities

  20. Use a defibrillator, save a life

    CERN Multimedia

    Joannah Caborn Wengler

    2012-01-01

    With the work for Long Shutdown 1 looming on the horizon, the CERN Fire Brigade is anticipating a heavy workload: more people working at CERN means more call-outs. So the more trained first-aiders around to help out before the paramedics arrive, the better. Would you know what to do in a medical emergency?   It could happen at any time: two colleagues are having a coffee at work, when one suddenly clutches his or her chest and falls to the floor unconscious. What would you do? Run to find a first-aider? Call the ambulance and wait, finishing your coffee? Neither response is entirely correct. On Monday 11 June in Building 40 the CMS safety group, in collaboration with the Fire Brigade and the Medical Service, demonstrated the recommended, potentially life-saving response to cardiac arrest (see the video), including the correct use of a defibrillator, ten of which were recently installed in key CERN locations (the Bulletin reported).     “In countries where...

  1. The challenges and possibilities of public access defibrillation

    DEFF Research Database (Denmark)

    Ringh, Mattias; Hollenberg, Jacob; Palsgaard-Moeller, Thea

    2018-01-01

    . Much hope has been put to the concept of Public Access Defibrillation with a massive dissemination of public available AEDs throughout most western countries. Accordingly, current guidelines recommend that AEDs should be deployed in places with a high likelihood of OHCA. Despite these efforts, AED use...... is in most settings anecdotal with little effect on overall OHCA survival. The major reasons for low use of public AEDs are that most OHCA take place outside high incidence sites of cardiac arrest and that most OHCAs take place in residential settings, currently defined as not suitable for Public Access...... Defibrillation. However, the use of new technology for identification and recruitment of lay bystanders and nearby AEDs to the scene of the cardiac arrest as well as new methods for strategic AED placement redefines and challenges the current concept and definitions of Public Access Defibrillation. Existing...

  2. Praehospital-hjertestopbehandling med semiautomatisk defibrillator--Heartstart 2000

    DEFF Research Database (Denmark)

    Fonsmark, L; Sandøe, E; Kastrup, J

    1989-01-01

    %. No practical problems of note occurred in connection with employment of the defibrillator. The ambulance staffs underwent six hours of training and this appeared to be adequate. It is concluded that Heartstart 2000 functions well and effectively in connection with revival of patients with cardiac arrest......In order to test the efficacy of a semiautomatic defibrillator (Heartstart 2000) in connection with cardiac arrest outside hospital, the apparatus was installed in two of the ambulances belonging to the Copenhagen Fire Service. The ambulance district involved was also equipped with an ambulance...... staffed by a doctor. A total of 48 patients with cardiac arrest were found and 16 of these had ventricular fibrillation. Six of the 16 patients have since been discharged from hospital (37.5%). The defibrillator had a high diagnostic certainty with a sensitivity of 96.5% and a specificity of 100...

  3. Praehospital-hjertestopbehandling med semiautomatisk defibrillator--Heartstart 2000

    DEFF Research Database (Denmark)

    Fonsmark, L; Sandøe, E; Kastrup, J

    1989-01-01

    staffed by a doctor. A total of 48 patients with cardiac arrest were found and 16 of these had ventricular fibrillation. Six of the 16 patients have since been discharged from hospital (37.5%). The defibrillator had a high diagnostic certainty with a sensitivity of 96.5% and a specificity of 100......%. No practical problems of note occurred in connection with employment of the defibrillator. The ambulance staffs underwent six hours of training and this appeared to be adequate. It is concluded that Heartstart 2000 functions well and effectively in connection with revival of patients with cardiac arrest...

  4. Automated cardiopulmonary resuscitation using a load-distributing band external cardiac support device for in-hospital cardiac arrest: a single centre experience of AutoPulse-CPR.

    Science.gov (United States)

    Spiro, J R; White, S; Quinn, N; Gubran, C J; Ludman, P F; Townend, J N; Doshi, S N

    2015-02-01

    Poor quality cardiopulmonary resuscitation (CPR) predicts adverse outcome. During invasive cardiac procedures automated-CPR (A-CPR) may help maintain effective resuscitation. The use of A-CPR following in-hospital cardiac arrest (IHCA) remains poorly described. Firstly, we aimed to assess the efficiency of healthcare staff using A-CPR in a cardiac arrest scenario at baseline, following re-training and over time (Scenario-based training). Secondly, we studied our clinical experience of A-CPR at our institution over a 2-year period, with particular emphasis on the details of invasive cardiac procedures performed, problems encountered, resuscitation rates and in-hospital outcome (AutoPulse-CPR Registry). Scenario-based training: Forty healthcare professionals were assessed. At baseline, time-to-position device was slow (mean 59 (±24) s (range 15-96s)), with the majority (57%) unable to mode-switch. Following re-training time-to-position reduced (28 (±9) s, pCPR Registry: 285 patients suffered IHCA, 25 received A-CPR. Survival to hospital discharge following conventional CPR was 28/260 (11%) and 7/25 (28%) following A-CPR. A-CPR supported invasive procedures in 9 patients, 2 of whom had A-CPR dependant circulation during transfer to the catheter lab. A-CPR may provide excellent haemodynamic support and facilitate simultaneous invasive cardiac procedures. A significant learning curve exists when integrating A-CPR into clinical practice. Further studies are required to better define the role and effectiveness of A-CPR following IHCA. Copyright © 2014 Elsevier Ireland Ltd. All rights reserved.

  5. Testing of Anesthesia Machines and Defibrillators in Healthcare Institutions.

    Science.gov (United States)

    Gurbeta, Lejla; Dzemic, Zijad; Bego, Tamer; Sejdic, Ervin; Badnjevic, Almir

    2017-09-01

    To improve the quality of patient treatment by improving the functionality of medical devices in healthcare institutions. To present the results of the safety and performance inspection of patient-relevant output parameters of anesthesia machines and defibrillators defined by legal metrology. This study covered 130 anesthesia machines and 161 defibrillators used in public and private healthcare institutions, during a period of two years. Testing procedures were carried out according to international standards and legal metrology legislative procedures in Bosnia and Herzegovina. The results show that in 13.84% of tested anesthesia machine and 14.91% of defibrillators device performance is not in accordance with requirements and should either have its results be verified, or the device removed from use or scheduled for corrective maintenance. Research emphasizes importance of independent safety and performance inspections, and gives recommendations for the frequency of inspection based on measurements. Results offer implications for adequacy of preventive and corrective maintenance performed in healthcare institutions. Based on collected data, the first digital electronical database of anesthesia machines and defibrillators used in healthcare institutions in Bosnia and Herzegovina is created. This database is a useful tool for tracking each device's performance over time.

  6. Worldwide experience with a totally subcutaneous implantable defibrillator

    DEFF Research Database (Denmark)

    Lambiase, Pier D; Barr, Craig; Theuns, Dominic A M J

    2014-01-01

    AIMS: The totally subcutaneous implantable-defibrillator (S-ICD) is a new alternative to the conventional transvenous ICD system to minimize intravascular lead complications. There are limited data describing the long-term performance of the S-ICD. This paper presents the first large internationa...

  7. Pacemakers and Implantable Defibrillators: MedlinePlus Health Topic

    Science.gov (United States)

    ... ClinicalTrials.gov: Pacemaker, Artificial (National Institutes of Health) Journal Articles References and abstracts from MEDLINE/PubMed (National ... Leadless Cardiac Pacemakers: The Next Evolution in Pacemaker Technology. ... on Pacemakers and Implantable Defibrillators is the National Heart, Lung, and Blood Institute Other Languages Find health information in languages other than English on Pacemakers and ...

  8. Implantable cardioverter defibrillator specific rehabilitation improves health cost outcomes

    DEFF Research Database (Denmark)

    Berg, Selina Kikkenborg; Zwisler, Ann-Dorthe; Koch, Mette Bjerrum

    2015-01-01

    OBJECTIVE: The Copenhagen Outpatient ProgrammE - implantable cardioverter defibrillator (COPE-ICD) trial included patients with implantable cardioverter defibrillators in a randomized controlled trial of rehabilitation. After 6-12 months significant differences were found in favour of the rehabil...... was -6,789 USD/-5,593 Euro in favour of rehabilitation. CONCLUSION: No long-term health outcome benefits were found for the rehabilitation programme. However, the rehabilitation programme resulted in a reduction in total attributable direct costs....... of the rehabilitation group for exercise capacity, general and mental health. The aim of this paper is to explore the long-term health effects and cost implications associated with the rehabilitation programme; more specifically, (i) to compare implantable cardioverter defibrillator therapy history and mortality...... between rehabilitation and usual care groups; (ii) to examine the difference between rehabilitation and usual care groups in terms of time to first admission; and (iii) to determine attributable direct costs. METHODS: Patients with first-time implantable cardioverter defibrillator implantation (n = 196...

  9. Risk of chronic anxiety in implantable defibrillator patients

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; den Broek, Krista C van; Theuns, Dominic A M J

    2011-01-01

    Little is known about the prevalence of chronic anxiety in patients with an implantable cardioverter defibrillator (ICD). In a multi-center, prospective study, we examined 1) the prevalence of chronic anxiety (i.e., patients anxious at implantation and 12 months), and 2) predictors of chronic...... anxiety....

  10. Daily remote monitoring of implantable cardioverter-defibrillators

    DEFF Research Database (Denmark)

    Hindricks, Gerhard; Varma, Niraj; Kacet, Salem

    2017-01-01

    Aims: Remote monitoring of implantable cardioverter-defibrillators may improve clinical outcome. A recent meta-analysis of three randomized controlled trials (TRUST, ECOST, IN-TIME) using a specific remote monitoring system with daily transmissions [Biotronik Home Monitoring (HM)] demonstrated...

  11. Shock whilst gardening--implantable defibrillators & lawn mowers.

    Science.gov (United States)

    Von Olshausen, G; Lennerz, C; Grebmer, C; Pavaci, H; Kolb, C

    2014-02-01

    Electromagnetic interference with implantable cardioverter defibrillators (ICDs) can cause inappropriate shock delivery or temporary inhibition of ICD functions. We present a case of electromagnetic interference between a lawn mower and an ICD resulting in an inappropriate discharge of the device due to erroneous detection of ventricular fibrillation.

  12. Externally Verifiable Oblivious RAM

    Directory of Open Access Journals (Sweden)

    Gancher Joshua

    2017-04-01

    Full Text Available We present the idea of externally verifiable oblivious RAM (ORAM. Our goal is to allow a client and server carrying out an ORAM protocol to have disputes adjudicated by a third party, allowing for the enforcement of penalties against an unreliable or malicious server. We give a security definition that guarantees protection not only against a malicious server but also against a client making false accusations. We then give modifications of the Path ORAM [15] and Ring ORAM [9] protocols that meet this security definition. These protocols both have the same asymptotic runtimes as the semi-honest original versions and require the external verifier to be involved only when the client or server deviates from the protocol. Finally, we implement externally verified ORAM, along with an automated cryptocurrency contract to use as the external verifier.

  13. Advancement of the methodology for automated integration of external hazards into level 1 PSA modeling. Technical report; Weiterentwicklung der Methodik zur automatisierten Integration uebergreifender Einwirkungen in PSA-Modelle der Stufe 1. Technischer Fachbericht

    Energy Technology Data Exchange (ETDEWEB)

    Berner, Nadine; Herb, Joachim

    2017-03-15

    In the course of the research and development project RS1539 funded by the German Federal Ministry for Economics and Energy (BMWi) the methodology for the automated integration of hazards in Level 1 PSA models has been enhanced. Thereby, the analysis tool pyRiskRobot provides the methodological framework for mapping a generic spectrum of internal and external hazards onto complex PSA plant models. The reimplementation of the software tool via the programming language python extends the applicability and facilitates the handling of pyRiskRobot in comparison to the previous Ruby-based version RiskRobot. Moreover, the development of functions to perform the topological modelling of fault trees and the probabilistic specification of modified fault tree elements have been continued. Due to the reimplementation and further developments, the tool enables to systematically generate fault trees of varying complexity, to flexibly integrate fault trees in existing PSA models and to automatically duplicate interconnected topologies. Thus, pyRiskRobot allows the efficient and traceable realization of hazard specific, usually laborious modifications of PSA models. In addition, pyRiskRobot has been extended to serve as a functional interface between the data compilations comprising the potential influences of hazards on PSA relevant components and the data base of a PSA plant model. Based on this conceptual design, additional analyses of the data can be carried out prior to the integration within the PSA model topology. The reimplemented functionalities of pyRiskRobot have been validated with respect to reference applications, such as the modelling of an internal fire scenario, against the previous version RiskRobot. The existing method collection for the automated modification of fault tree topologies has been extended based on the requirements for further applications, among others the modelling of an external flooding scenario. The deduced hazard specific modelling approaches

  14. Resins production: batch plant automation

    International Nuclear Information System (INIS)

    Banti, M.; Mauri, G.

    1996-01-01

    Companies that look for automation in their plants without external resources, have at their disposal flexible, custom and easy to use DCS, open towards PLC. In this article it is explained why Hoechts has followed this way of new plants for resins production automation

  15. Factors associated with delayed defibrillation in cardiopulmonary resuscitation: A prospective simulation study.

    Directory of Open Access Journals (Sweden)

    Christoph Castan

    Full Text Available Early defibrillation is an important factor of survival in cardiac arrest. However, novice resuscitators often struggle with cardiac arrest patients. We investigated factors leading to delayed defibrillation performed by final-year medical students within a simulated bystander cardiac arrest situation.Final-year medical students received a refresher lecture and basic life support training before being confronted with a simulated cardiac arrest situation in a simulation ambulance. The scenario was analyzed for factors leading to delayed defibrillation. We compared the time intervals the participants needed for various measures with a benchmark set by experienced resuscitators. After training, the participants were interviewed regarding challenges and thoughts during the scenario.The median time needed for defibrillation was 158 s (n = 49, interquartile range: 107-270 s, more than six-fold of the benchmark time. The major part of total defibrillation time (49%; median, n = 49 was between onset of ventricular fibrillation and beginning to prepare the defibrillator, more specifically the time between end of preparation of the defibrillator and actual delivery of the shock, with a mean proportion of 26% (n = 49, SD = 17% of the overall time needed for defibrillation (maximum 67%. Self-reported reasons for this delay included uncertainty about the next step to take, as reported by 73% of the participants. A total of 35% were unsure about which algorithm to follow. Diagnosing the patient was subjectively difficult for 35% of the participants. Overall, 53% of the participants felt generally confused.Our study shows that novice resuscitators rarely achieve guideline-recommended defibrillation times. The most relative delays were observed when participants had to choose what to do next or which algorithm to follow, and thus i.e. performed extensive airway management before a life-saving defibrillation. Our data provides a first insight in the process of

  16. Live defibrillation in simulation-based medical education--a survey of simulation center practices and attitudes.

    Science.gov (United States)

    Turban, Joseph W; Peters, Deborah P; Berg, Benjamin W

    2010-02-01

    Resuscitation from cardiac arrhythmia, requiring cardioversion/defibrillation is a common simulation training scenario. Use of live defibrillation enhances simulation fidelity but is not without risk. This survey was conducted to describe the prevalence of live defibrillation use during training scenarios in healthcare simulation centers, and when used, if safety training was required before using live defibrillation. A convenience sample of attendees at the 7th annual International Meeting on Simulation in Healthcare (January 2007) was surveyed using a closed-ended 23-item survey instrument. Survey domains included responder and simulation center demographics, simulation center defibrillation safety policies, and attitudes toward defibrillation practices in simulation training environments. Fifty-seven individuals representing 39 simulation centers returned surveys, 29 of which were in the United States. Live defibrillation was used in 35 of the 39 centers (90%). A defibrillation safety training policy was in effect at 14 of 39 centers (36%). Formal training before using live defibrillation was considered necessary by 48 of 55 responders (87%). Forty-eight of 54 responders (89%) strongly agreed or agreed with the statement, "I feel using live defibrillation plays an important role in simulation-based education." Although most responders consider use of live defibrillation important and believe formal defibrillator safety training should be conducted before use, only about one third of the centers had a training policy in effect. It remains to be determined whether safety training before the use of live defibrillation during simulation-based education increases user safety.

  17. The Effects of Normothermic and Hypothermic Cardiopulmonary Bypass Upon Defibrillation Energy Requirements and Transmyocardial Impedance

    National Research Council Canada - National Science Library

    Martin, David

    1993-01-01

    .... To evaluate these questions we studied the effect of controlled hypothermia upon defibrillation energy requirements and transcardiac impedance in a canine model of cardiopulmonary bypass in which 26...

  18. Management of Implantable Cardioverter Defibrillator Recipients: Care Beyond Guidelines.

    Science.gov (United States)

    Philippon, François; Sterns, Laurence D; Nery, Pablo B; Parkash, Ratika; Birnie, David; Rinne, Claus; Mondesert, Blandine; Exner, Derek; Bennett, Matthew

    2017-08-01

    This companion article is intended to address common clinical scenarios in patients with implantable defibrillators that were not addressed in the 2016 Canadian Cardiovascular Society/Canadian Heart Rhythm Society implantable cardioverter defibrillator guidelines including recommendations for device programming to improve detection, to minimize shocks (appropriate and inappropriate), and to minimize ventricular pacing. Important issues at the time of replacement such as device prescription, technical aspects (vascular access, extraction), and management of components on advisories are also discussed. Finally, common clinical scenarios such as management of patients with terminal illnesses, recurrent ventricular tachycardia, electrical storms, catheter ablation for ventricular tachycardia, and system infection management are considered. The management of these patients requires a team approach and comprehensive knowledge surrounding these common clinical scenarios. Copyright © 2017 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.

  19. Cardioverter-defibrillator implantation in myeloma-associated cardiac amyloidosis.

    Science.gov (United States)

    Campanile, Alfonso; Sozzi, Fabiola B; Canetta, Ciro; Danzi, Gian Battista

    2013-01-01

    A 62-year-old woman with multiple myeloma and light-chain amyloidosis with significant heart involvement developed an in-hospital cardiac arrest. After cardiopulmonary resuscitation, a stable sinus rhythm without any cerebral damage was restored, and the patient was admitted to the coronary care unit. A cardioverter-defibrillator was implanted, and it successfully intervened in two sustained ventricular tachycardia episodes and one ventricular fibrillation episode, which were recorded during hospitalization. After achieving discrete cardiac compensation, the patient was transferred to the emergency medicine department where she underwent chemotherapy for multiple myeloma. The patient died 40 days after admission from refractory heart failure. In the literature, there are studies that describe the use of cardioverter-defibrillator implantation in cardiac amyloidosis; however, at present, there is no evidence of a beneficial effect on survival with the use of this intervention. A high index of suspicion for amyloid heart disease and early diagnosis are critical to improving outcomes.

  20. Behavioral interventions in patients with an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Habibović, Mirela; Burg, Matthew M; Pedersen, Susanne S.

    2013-01-01

    The implantable cardioverter defibrillator (ICD) is the first-line treatment for primary and secondary prevention of sudden cardiac death. A subgroup of patients experience psychological distress postimplant, and no clear evidence base exists regarding how best to address patients' needs. The aim...... of this critical review is to provide an overview of behavioral interventions in ICD patients to date, and to delineate directions for future research using lessons learned from the ongoing RISTA and WEBCARE trials....

  1. The NO Regular Defibrillation testing In Cardioverter Defibrillator Implantation (NORDIC ICD) trial: concept and design of a randomized, controlled trial of intra-operative defibrillation testing during de novo defibrillator implantation.

    Science.gov (United States)

    Bänsch, Dietmar; Bonnemeier, Hendrik; Brandt, Johan; Bode, Frank; Svendsen, Jesper Hastrup; Felk, Angelika; Hauser, Tino; Wegscheider, Karl

    2015-01-01

    Although defibrillation (DF) testing is still considered a standard procedure during implantable cardioverter-defibrillator (ICD) insertion and has been an essential element of all trials that demonstrated the survival benefit of ICD therapy, there are no large randomized clinical trials demonstrating that DF testing improves clinical outcome and if the outcome would remain the same by omitting DF testing. Between February 2011 and July 2013, we randomly assigned 1077 patients to ICD implantation with (n = 540) or without (n = 537) DF testing. The intra-operative DF testing was standardized across all participating centres. After inducing a fast ventricular tachycardia (VT) with a heart rate ≥240 b.p.m. or ventricular fibrillation (VF) with a low-energy T-wave shock, DF was attempted with an initial 15 J shock. If the shock reversed the VT or VF, DF testing was considered successful and terminated. If unsuccessful, two effective 24 J shocks were administered. If DF was unsuccessful, the system was reconfigured and another DF testing was performed. An ICD shock energy of 40 J had to be programmed in all patients for treatment of spontaneous VT/VF episodes. The primary endpoint was the average efficacy of the first ICD shock for all true VT/VF episodes in each patient during follow-up. The secondary endpoints included the frequency of system revisions, total fluoroscopy, implantation time, procedural serious adverse events, and all-cause, cardiac, and arrhythmic mortality during follow-up. Home Monitoring was used in all patients to continuously monitor the system integrity, device programming and performance. The NO Regular Defibrillation testing In Cardioverter Defibrillator Implantation (NORDIC ICD) trial is one of two large prospective randomized trials assessing the effect of DF testing omission during ICD implantation. NCT01282918. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email

  2. Law Enforcement Agency Defibrillation (LEA-D): proceedings of the National Center for Early Defibrillation Police AED Issues Forum.

    Science.gov (United States)

    Mosesso, Vincent N; Newman, Mary M; Ornato, Joseph P; Paris, Paul M; Andersen, Leon; Brinsfield, Kathryn; Dunnavant, Gregory R; Frederick, Jay; Groh, William J; Johnston, Steven; Lerner, E Brooke; Murphy, George P; Myerburg, Robert J; Rosenberg, Donald G; Savino, Mitchell; Sayre, Michael R; Sciammarella, Joseph; Schoen, Valerie; Vargo, Philip; van Alem, Anouk; White, Roger D

    2002-01-01

    Why does LEA-D intervention seem to work in some systems but not others? Panelists agreed that some factors that delay rapid access to treatment, such as long travel distances in rural areas, may represent insurmountable barriers. Other factors, however, may be addressed more readily. These include: absence of a medical response culture, discomfort with the role of medical intervention, insecurity with the use of medical devices, a lack of proactive medical direction, infrequent refresher training, and dependence on EMS intervention. Panelists agreed that successful LEA-D programs possess ten key attributes (Table 6). In the end, the goal remains "early" defibrillation, not "police" defibrillation. It does not matter whether the rescuer wears a blue uniform--or any uniform, for that matter--so long as the defibrillator reaches the victim quickly. If LEA personnel routinely arrive at medical emergencies after other emergency responders or after 8 minutes have elapsed from the time of collapse, an LEA-D program will be unlikely to provide added value. Similarly, if police frequently arrive first, but the department is unwilling or unable to cultivate the attributes of successful LEA-D programs, efforts to improve survival may not be realized. In most communities, however, LEA-D programs have tremendous lifesaving potential and are well worth the investment of time and resources. Law enforcement agencies considering adoption of AED programs should review the frequency with which police arrive first at medical emergencies and LEA response intervals to determine whether AED programs might help improve survival in their communities. It is time for law enforcement agency defibrillation to become the rule, not the exception.

  3. Effect of defibrillation threshold testing on effectiveness of the subcutaneous implantable cardioverter defibrillator.

    Science.gov (United States)

    Peddareddy, Lakshmi; Merchant, Faisal M; Leon, Angel R; Smith, Paige; Patel, Akshar; El-Chami, Mikhael F

    2018-06-12

    Defibrillation threshold (DFT) testing is recommended with the subcutaneous ICD (SICD). To describe first shock efficacy for appropriate SICD therapies stratified by the presence of implant DFT testing. We reviewed all patients receiving SICDs at our institution and stratified them based on whether implant DFT testing was performed. Appropriate shocks were reviewed to see if ventricular tachycardia/ventricular fibrillation (VT/VF) terminated with a single shock. First shock efficacy was stratified by implant DFT status. 178 patients implanted with SICDs and followed in our center were included in this study. Of these, 135 (76 %) underwent DFT testing (DFT (+) group). In the DFT (+) 80 appropriate shocks were needed to treat 69 episodes of VT/VF. The first shock was effective in 61 out of 69 episodes (88.4 %), whereas multiple shocks were required to terminate VT/VF in the remaining 8 episodes. Among 43 patients without implant DFT testing (DFT (-) group), 20 appropriate shocks to treat 17 episodes of VT/VF occurred in 7 patients. VT/VF was successfully terminated with the first shock in 16 out of 17 episodes (first shock efficacy 94.1 %). There was no significant difference in first shock effectiveness between those with and without implant DFT testing (p = 0.97). A strategy that omits DFT testing at implant did not appear to compromise the effictiveness of the SICD. These data suggest that routine DFT testing at SICD implant might not be necessary. Randomized trials are needed to confirm this finding. This article is protected by copyright. All rights reserved. This article is protected by copyright. All rights reserved.

  4. Suppression of AC railway power-line interference in ECG signals recorded by public access defibrillators

    Directory of Open Access Journals (Sweden)

    Dotsinsky Ivan

    2005-11-01

    Full Text Available Abstract Background Public access defibrillators (PADs are now available for more efficient and rapid treatment of out-of-hospital sudden cardiac arrest. PADs are used normally by untrained people on the streets and in sports centers, airports, and other public areas. Therefore, automated detection of ventricular fibrillation, or its exclusion, is of high importance. A special case exists at railway stations, where electric power-line frequency interference is significant. Many countries, especially in Europe, use 16.7 Hz AC power, which introduces high level frequency-varying interference that may compromise fibrillation detection. Method Moving signal averaging is often used for 50/60 Hz interference suppression if its effect on the ECG spectrum has little importance (no morphological analysis is performed. This approach may be also applied to the railway situation, if the interference frequency is continuously detected so as to synchronize the analog-to-digital conversion (ADC for introducing variable inter-sample intervals. A better solution consists of rated ADC, software frequency measuring, internal irregular re-sampling according to the interference frequency, and a moving average over a constant sample number, followed by regular back re-sampling. Results The proposed method leads to a total railway interference cancellation, together with suppression of inherent noise, while the peak amplitudes of some sharp complexes are reduced. This reduction has negligible effect on accurate fibrillation detection. Conclusion The method is developed in the MATLAB environment and represents a useful tool for real time railway interference suppression.

  5. Suppression of AC railway power-line interference in ECG signals recorded by public access defibrillators.

    Science.gov (United States)

    Dotsinsky, Ivan

    2005-11-26

    Public access defibrillators (PADs) are now available for more efficient and rapid treatment of out-of-hospital sudden cardiac arrest. PADs are used normally by untrained people on the streets and in sports centers, airports, and other public areas. Therefore, automated detection of ventricular fibrillation, or its exclusion, is of high importance. A special case exists at railway stations, where electric power-line frequency interference is significant. Many countries, especially in Europe, use 16.7 Hz AC power, which introduces high level frequency-varying interference that may compromise fibrillation detection. Moving signal averaging is often used for 50/60 Hz interference suppression if its effect on the ECG spectrum has little importance (no morphological analysis is performed). This approach may be also applied to the railway situation, if the interference frequency is continuously detected so as to synchronize the analog-to-digital conversion (ADC) for introducing variable inter-sample intervals. A better solution consists of rated ADC, software frequency measuring, internal irregular re-sampling according to the interference frequency, and a moving average over a constant sample number, followed by regular back re-sampling. The proposed method leads to a total railway interference cancellation, together with suppression of inherent noise, while the peak amplitudes of some sharp complexes are reduced. This reduction has negligible effect on accurate fibrillation detection. The method is developed in the MATLAB environment and represents a useful tool for real time railway interference suppression.

  6. Pre-implantation implantable cardioverter defibrillator concerns and Type D personality increase the risk of mortality in patients with an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; van den Broek, Krista C; Erdman, Ruud A M

    2010-01-01

    Little is known about the influence of psychological factors on prognosis in implantable cardioverter defibrillator (ICD) patients. We examined the influence of the distressed personality (Type D) and pre-implantation device concerns on short-term mortality in ICD patients.......Little is known about the influence of psychological factors on prognosis in implantable cardioverter defibrillator (ICD) patients. We examined the influence of the distressed personality (Type D) and pre-implantation device concerns on short-term mortality in ICD patients....

  7. Home Automation

    OpenAIRE

    Ahmed, Zeeshan

    2010-01-01

    In this paper I briefly discuss the importance of home automation system. Going in to the details I briefly present a real time designed and implemented software and hardware oriented house automation research project, capable of automating house's electricity and providing a security system to detect the presence of unexpected behavior.

  8. Depressive symptoms in patients with an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Pedersen, Susanne S; Andersen, Christina M; Denollet, Johan

    2018-01-01

    OBJECTIVE: Patients with an implantable cardioverter defibrillator (ICD) and co-morbid depression are at greater risk of poor quality of life and premature death. We examined if treatment expectations predict depressive symptoms 12months post implant. METHODS: First-time implant patients from...... of 12-months depressive symptoms: Model 1: Negative treatment expectations (β=0.202; p=0.020) and baseline depression (β=0.376; pdepression (β=0.350; p....051). Model 3: Baseline depression (β=0.353; p

  9. Making post-mortem implantable cardioverter defibrillator explantation safe

    DEFF Research Database (Denmark)

    Räder, Sune B E W; Zeijlemaker, Volkert; Pehrson, Steen

    2009-01-01

    that the resting voltage over the operating person would not exceed 50 V. CONCLUSION: The use of intact medical gloves made of latex, neoprene, or plastic eliminates the potential electrical risk during explantation of an ICD. Two gloves on each hand offer sufficient protection. We will recommend the use......AIMS: The aim of this study is to investigate whether protection with rubber or plastic gloves during post-mortem explantation of an implantable cardioverter defibrillator (ICD) offers enough protection for the explanting operator during a worst-case scenario (i.e. ICD shock). METHODS AND RESULTS...

  10. Peer-led training in basic life support and resuscitation using an automatic external defibrillator

    DEFF Research Database (Denmark)

    Løfgren, Bo; Petersen, Christina Børlum; Mikkelsen, Ronni

    2009-01-01

    Peer-led training has been identified as a useful tool for delivering undergraduate healthcare training. In this paper we describe the implementation of the European Resuscitation Council BLS/AED Course as a peer-led training program for medical students....

  11. [Electrical storm in patients with prophylactic defibrillator implantation].

    Science.gov (United States)

    Rodríguez-Mañero, Moisés; González-Cambeiro, Cristina; Moreno-Arribas, Jose; Expósito-García, Víctor; Sánchez-Gómez, Juan Miguel; González-Torres, Luis; Arce-León, Álvaro; Arguedas-Jiménez, Hugo; Gaztañaga, Larraitz; Salvador-Montañés, Oscar; Iglesias-Bravo, Jose Antonio; Huerta, Ana Andrés La; Fernández-Armenta, Juan; Arias, Miguel Ángel; Martínez-Sande, Luis

    2016-01-01

    Little is known about the prevalence of electrical storm, baseline characteristics and mortality implications of patients with implantable cardioverter defibrillator in primary prevention versus those patients without electrical storm. We sought to assess the prevalence, baseline risk profile and survival significance of electrical storm in patients with implantable defibrillator for primary prevention. Retrospective multicenter study performed in 15 Spanish hospitals. Consecutives patients referred for desfibrillator implantation, with or without left ventricular lead (at least those performed in 2010 and 2011), were included. Over all 1,174 patients, 34 (2,9%) presented an electrical storm, mainly due to ventricular tachycardia (82.4%). There were no significant baseline differences between groups, with similar punctuation in the mortality risk scores (SHOCKED, MADIT and FADES). A clear trigger was identified in 47% of the events. During the study period (38±21 months), long-term total mortality (58.8% versus 14.4%, pstorm patients. Rate of inappropriate desfibrillator intervention was also higher (14.7 versus 8.6%, pstorm was 2.9%. There were no baseline differences in the cardiovascular risk profile versus those without electrical storm. However, all cause mortality and cardiovascular mortality was increased in these patients versus control desfibrillator patients without electrical storm, as was the rate of inappropriate desfibrillator intervention. Copyright © 2015 Instituto Nacional de Cardiología Ignacio Chávez. Published by Masson Doyma México S.A. All rights reserved.

  12. Effects of postshock atrial pacing on atrial defibrillation outcome in the isolated sheep heart

    NARCIS (Netherlands)

    Skanes, A. C.; Gray, R. A.; Zuur, C. L.; Jalife, J.

    1998-01-01

    BACKGROUND: Failed atrial defibrillation shocks are associated with organization of postshock activity and a substantial postshock electrical quiescence. We investigated the ability of a train of pacing stimuli to capture or locally entrain atrial myocardium during the quiescent period after

  13. Somatosensory amplification mediates sex differences in psychological distress among cardioverter-defibrillator patients

    DEFF Research Database (Denmark)

    Versteeg, Henneke; Baumert, Jens; Kolb, Christof

    2010-01-01

    The present study examined whether female patients with an implantable cardioverter defibrillator (ICD) report more psychological distress than male patients, and whether somatosensory amplification mediates this relationship. Design: Consecutive ICD patients (N = 241; 33% women) participating in...

  14. Correlation of geomagnetic activity with implantable cardioverter defibrillator shocks and antitachycardia pacing

    Czech Academy of Sciences Publication Activity Database

    Ebrille, E.; Konecny, T.; Konecny, D.; Špaček, R.; Jones, P.; Ambrož, Pavel; DeSimone, C.V.; Powel, B.D.; Hayes, D.L.; Friedman, P.A.; Asirvatham, S.J.

    2015-01-01

    Roč. 90, č. 2 (2015), s. 202-208 ISSN 0025-6196 Institutional support: RVO:67985815 Keywords : geomagnetic activity * implantable cardioverter defibrillator Subject RIV: FA - Cardiovascular Diseases incl. Cardiotharic Surgery Impact factor: 5.920, year: 2015

  15. Tricuspid Valve Dysfunction Following Pacemaker or Cardioverter-Defibrillator Implantation.

    Science.gov (United States)

    Chang, James D; Manning, Warren J; Ebrille, Elisa; Zimetbaum, Peter J

    2017-05-09

    The potential for cardiac implantable electronic device leads to interfere with tricuspid valve (TV) function has gained increasing recognition as having hemodynamic and clinical consequences associated with incremental morbidity and death. The diagnosis and treatment of lead-related (as distinct from functional) tricuspid regurgitation pose unique challenges. Because of pitfalls in routine diagnostic imaging, a high level of clinical suspicion must be maintained to avoid overlooking the possibility that worsening heart failure is a consequence of mechanical interference with TV leaflet mobility or coaptation and is amenable to lead extraction or valve repair or replacement. The future of cardiac implantable electronic devices includes pacing and perhaps defibrillation without a lead traversing the TV. Copyright © 2017 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.

  16. Driving safety among patients with automatic implantable cardioverter defibrillators.

    Science.gov (United States)

    Finch, N J; Leman, R B; Kratz, J M; Gillette, P C

    1993-10-06

    To determine the driving behavior of patients following the placement of automatic implantable cardioverter defibrillators (AICDs). Forty patients with AICDs (33 men, seven women; mean age, 62.7 years) responded to a questionnaire designed to ascertain driving behavior after hospital discharge. Despite medical advice never to drive again, 28 patients (70%) resumed driving, with the majority doing so by 8 months after AICD implantation. Of these, 11 (40%) identified themselves as the primary driver in their household. Fourteen (50%) drove daily. Two (7%) were driving and continued to drive during discharge of their AICDs. Twenty-five (91%) reported that they felt comfortable and safe while driving. A majority of patients with AICDs continue to drive after a proscription of this activity by health care workers.

  17. Psychological distress in patients with an implantable cardioverter defibrillator

    Directory of Open Access Journals (Sweden)

    Tsuyoshi Shiga, MD

    2013-12-01

    Full Text Available Despite the effectiveness of an implantable cardioverter defibrillator (ICD, its implantation and concomitant therapies, including shocks, can induce psychological distress in patients. Depression has been observed in approximately 30% of ICD patients, and shocks may contribute to the persistence of depression. Anxiety is common, with reports of 24–87% of ICD patients experiencing symptoms of anxiety after implantation, and type D personality and ICD-related concerns may play important roles in the level of anxiety in ICD patients. However, the association between ICD shocks and anxiety is controversial. The prevalence of posttraumatic stress disorder (PTSD in ICD patients is approximately 20%, and type D personality, comorbidities, and frequent shocks may contribute to PTSD. It is also important to pay attention to the psychological distress in the partners of ICD patients.

  18. Defibrillator Implantation in Patients with Nonischemic Systolic Heart Failure

    DEFF Research Database (Denmark)

    Køber, Lars; Thune, Jens J; Nielsen, Jens C

    2016-01-01

    Background The benefit of an implantable cardioverter-defibrillator (ICD) in patients with symptomatic systolic heart failure caused by coronary artery disease has been well documented. However, the evidence for a benefit of prophylactic ICDs in patients with systolic heart failure that is not due...... to coronary artery disease has been based primarily on subgroup analyses. The management of heart failure has improved since the landmark ICD trials, and many patients now receive cardiac resynchronization therapy (CRT). Methods In a randomized, controlled trial, 556 patients with symptomatic systolic heart.......6%) in the control group (P=0.29). Conclusions In this trial, prophylactic ICD implantation in patients with symptomatic systolic heart failure not caused by coronary artery disease was not associated with a significantly lower long-term rate of death from any cause than was usual clinical care. (Funded by Medtronic...

  19. Pacing and Defibrillators in Complex Congenital Heart Disease

    Science.gov (United States)

    Chubb, Henry; O’Neill, Mark; Rosenthal, Eric

    2016-01-01

    Device therapy in the complex congenital heart disease (CHD) population is a challenging field. There is a myriad of devices available, but none designed specifically for the CHD patient group, and a scarcity of prospective studies to guide best practice. Baseline cardiac anatomy, prior surgical and interventional procedures, existing tachyarrhythmias and the requirement for future intervention all play a substantial role in decision making. For both pacing systems and implantable cardioverter defibrillators, numerous factors impact on the merits of system location (endovascular versus non-endovascular), lead positioning, device selection and device programming. For those with Fontan circulation and following the atrial switch procedure there are also very specific considerations regarding access and potential complications. This review discusses the published guidelines, device indications and the best available evidence for guidance of device implantation in the complex CHD population. PMID:27403295

  20. Effects of advanced life support on patients who suffered cardiac arrest outside of hospital and were defibrillated.

    Science.gov (United States)

    Hagihara, Akihito; Onozuka, Daisuke; Nagata, Takashi; Hasegawa, Manabu

    2018-01-01

    The effects and relative benefits of advanced airway management and epinephrine on patients with out-of-hospital cardiac arrest (OHCA) who were defibrillated are not well understood. This was a prospective observational study. Using data of all out-of-hospital cardiac arrest cases occurring between 2005 and 2013 in Japan, hierarchical logistic regression and conditional logistic regression along with time-dependent propensity matching were performed. Outcome measures were survival and minimal neurological impairment [cerebral performance category (CPC) 1 or 2] at 1month after the event. We analyzed 37,873 cases that met the inclusion criteria. Among propensity-matched patients, advanced airway management and/or prehospital epinephrine use was related to decreased rates of 1-month survival (adjusted odds ratio 0.88, 95% confidence interval 0.80 to 0.97) and CPC (1, 2) (adjusted odds ratio 0.56, 95% confidence interval 0.48 to 0.66). Advanced airway management was related to decreased rates of 1-month survival (adjusted odds ratio 0.89, 95% confidence interval 0.81to 0.98) and CPC (1, 2) (adjusted odds ratio 0.54, 95% confidence interval 0.46 to 0.64) in patients who did not receive epinephrine, whereas epinephrine use was not related to the outcome measures. In defibrillated patients with OHCA, advanced airway management and/or epinephrine are related to reduced long-term survival, and advanced airway management is less beneficial than epinephrine. However, the proportion of patients with OHCA who responded to an initial shock was very low in the study subjects, and the external validity of our findings might be limited. Copyright © 2017 Elsevier Inc. All rights reserved.

  1. A Study on Performance and Safety Tests of Defibrillator Equipment

    Directory of Open Access Journals (Sweden)

    Tavakoli Golpaygani A.

    2017-12-01

    Full Text Available Introduction: Nowadays, more than 10,000 different types of medical devices can be found in hospitals. This way, medical electrical equipment is being employed in a wide variety of fields in medical sciences with different physiological effects and measurements. Hospitals and medical centers must ensure that their critical medical devices are safe, accurate, reliable and operational at the required level of performance. Defibrillators are critical resuscitation devices. The use of reliable defibirillators has led to more effective treatments and improved patient safety through better control and management of complications during Cardiopulmonary Resuscitation (CPR. Materials and Methods: The metrological reliability of twenty frequent use, manual defibrillators in use ten hospitals (4 private and 6 public in one of the provinces of Iran according to international and national standards was evaluated. Results: Quantitative analysis of control and instrument accuracy showed the amount of the obtained results in many units are critical which had less value over the standard limitations especially in devices with poor battery. For the accuracy of delivered energy analysis, only twelve units delivered acceptable output values and the precision in the output energy measurements especialy in weak battry condition, after activation of discharge alarm, were low. Conclusion: Obtained results indicate a need for new and severe regulations on periodic performance verifications and medical equipment quality control program especially for high risk instruments. It is also necessary to provide training courses on the fundumentals of operation and performane parameters for medical staff in the field of meterology in medicine and how one can get good accuracy results especially in high risk medical devices.

  2. A Study on Performance and Safety Tests of Defibrillator Equipment.

    Science.gov (United States)

    Tavakoli Golpaygani, A; Movahedi, M M; Reza, M

    2017-12-01

    Nowadays, more than 10,000 different types of medical devices can be found in hospitals. This way, medical electrical equipment is being employed in a wide variety of fields in medical sciences with different physiological effects and measurements. Hospitals and medical centers must ensure that their critical medical devices are safe, accurate, reliable and operational at the required level of performance. Defibrillators are critical resuscitation devices. The use of reliable defibirillators has led to more effective treatments and improved patient safety through better control and management of complications during Cardiopulmonary Resuscitation (CPR). The metrological reliability of twenty frequent use, manual defibrillators in use ten hospitals (4 private and 6 public) in one of the provinces of Iran according to international and national standards was evaluated. Quantitative analysis of control and instrument accuracy showed the amount of the obtained results in many units are critical which had less value over the standard limitations especially in devices with poor battery. For the accuracy of delivered energy analysis, only twelve units delivered acceptable output values and the precision in the output energy measurements especialy in weak battry condition, after activation of discharge alarm, were low. Obtained results indicate a need for new and severe regulations on periodic performance verifications and medical equipment quality control program especially for high risk instruments. It is also necessary to provide training courses on the fundumentals of operation and performane parameters for medical staff in the field of meterology in medicine and how one can get good accuracy results especially in high risk medical devices.

  3. Clinical performance of different DF-4 implantable cardioverter defibrillator leads.

    Science.gov (United States)

    Sarrazin, Jean-François; Philippon, François; Sellier, Romain; André, Philippe; O'Hara, Gilles; Molin, Franck; Nault, Isabelle; Blier, Louis; Champagne, Jean

    2018-06-01

    Implantable cardioverter-defibrillator (ICD) DF-4 connectors have been introduced to facilitate defibrillator lead connection and to reduce the size of device header. There are limited data regarding the overall performance of those leads and no comparison between different ICD DF-4 leads. This is a cohort study of consecutive patients implanted with ICD DF-4 lead system at one University Centre between October 2010 and February 2015. A historical control group of patients with ICD DF-1 lead implantation was used for comparison. The following ICD DF-4 leads were evaluated: St. Jude Medical Durata 7122Q (St. Jude Medical, St. Paul, MN, USA), Medtronic Sprint Quattro Secure 6935 M (Medtronic Inc., Minneapolis, MN, USA), Boston Scientific Endotak Reliance 4-Site 0293 (Boston Scientific, Marlborough, MA, USA), and Boston Scientific Reliance 4-Front 0693. This study evaluated the acute and mid-term performances of those leads as well as complications. A total of 812 patients (age 63 ± 12 years, 80% male, left ventricular ejection fraction 31 ± 12%) underwent implantation of an ICD DF-4 lead. Acute and follow-up R-wave sensing and threshold were excellent. Compared to implantation, intrinsic R waves were higher at follow-up for Boston Scientific and Medtronic leads, and pacing lead impedances were lower for all leads at first follow-up (P < 0.001). The number of lead dislodgement or failure was similar between all leads. The estimated lead survival rates at 3 years were 95.6% for Boston Scientific Endotak 4-Site, 97.1% for Boston Scientific 4-Front, 97.7% for Medtronic Sprint Quattro, and 97.5% for St. Jude Durata (P  =  0.553). All ICD DF-4 leads had excellent acute and mid-term electrical performances. Longer follow-up will be necessary to confirm their sustained performance. © 2018 Wiley Periodicals, Inc.

  4. Hands-On Defibrillation Skills of Pediatric Acute Care Providers During a Simulated Ventricular Fibrillation Cardiac Arrest Scenario.

    Science.gov (United States)

    Bhalala, Utpal S; Balakumar, Niveditha; Zamora, Maria; Appachi, Elumalai

    2018-01-01

    Introduction: Timely defibrillation in ventricular fibrillation cardiac arrest (VFCA) is associated with good outcome. While defibrillation skills of pediatric providers have been reported to be poor, the factors related to poor hands-on defibrillation skills of pediatric providers are largely unknown. The aim of our study was to evaluate delay in individual steps of the defibrillation and human and non-human factors associated with poor hands-on defibrillation skills among pediatric acute care providers during a simulated VFCA scenario. Methods: We conducted a prospective observational study of video evaluation of hands-on defibrillation skills of pediatric providers in a simulated VFCA in our children's hospital. Each provider was asked to use pads followed by paddles to provide 2 J/kg shock to an infant mannequin in VFCA. The hands-on skills were evaluated for struggle with any step of defibrillation, defined a priori as >10 s delay with particular step. The data was analyzed using chi-square test with significant p -value 10 s delay) with each of connecting the pads/paddles to the device, using pads/paddles on the mannequin and using buttons on the machine was 34 (50%), 26 (38%), and 31 (46%), respectively. Conclusions: The defibrillation skills of providers in a tertiary care children's hospital are poor. Both human and machine-related factors are associated with delay in defibrillation. Prior use of the study defibrillator is associated with a significantly shorter time-to-first shock as compared to prior use of any other defibrillator or no prior use of any defibrillator.

  5. Process automation

    International Nuclear Information System (INIS)

    Moser, D.R.

    1986-01-01

    Process automation technology has been pursued in the chemical processing industries and to a very limited extent in nuclear fuel reprocessing. Its effective use has been restricted in the past by the lack of diverse and reliable process instrumentation and the unavailability of sophisticated software designed for process control. The Integrated Equipment Test (IET) facility was developed by the Consolidated Fuel Reprocessing Program (CFRP) in part to demonstrate new concepts for control of advanced nuclear fuel reprocessing plants. A demonstration of fuel reprocessing equipment automation using advanced instrumentation and a modern, microprocessor-based control system is nearing completion in the facility. This facility provides for the synergistic testing of all chemical process features of a prototypical fuel reprocessing plant that can be attained with unirradiated uranium-bearing feed materials. The unique equipment and mission of the IET facility make it an ideal test bed for automation studies. This effort will provide for the demonstration of the plant automation concept and for the development of techniques for similar applications in a full-scale plant. A set of preliminary recommendations for implementing process automation has been compiled. Some of these concepts are not generally recognized or accepted. The automation work now under way in the IET facility should be useful to others in helping avoid costly mistakes because of the underutilization or misapplication of process automation. 6 figs

  6. Distribution automation

    International Nuclear Information System (INIS)

    Gruenemeyer, D.

    1991-01-01

    This paper reports on a Distribution Automation (DA) System enhances the efficiency and productivity of a utility. It also provides intangible benefits such as improved public image and market advantages. A utility should evaluate the benefits and costs of such a system before committing funds. The expenditure for distribution automation is economical when justified by the deferral of a capacity increase, a decrease in peak power demand, or a reduction in O and M requirements

  7. Sustaining cyborgs: sensing and tuning agencies of pacemakers and implantable cardioverter defibrillators.

    Science.gov (United States)

    Oudshoorn, Nelly

    2015-02-01

    Recently there has been a renewed interest in cyborgs, and particularly in new and emerging fusions of humans and technologies related to the development of human enhancement technologies. These studies reflect a trend to follow new and emerging technologies. In this article, I argue that it is important to study 'older' and more familiar cyborgs as well. Studying 'the old' is important because it enables us to recognize hybrids' embodied experiences. This article addresses two of these older hybrids: pacemakers and implantable cardioverter defibrillators inserted in the bodies of people suffering from heart-rhythm disturbances. My concern with hybrid bodies is that internal devices seem to present a complex and neglected case if we wish to understand human agency. Their 'users' seem to be passive because they cannot exert any direct control over the working of their devices. Technologies inside bodies challenge a longstanding tradition of theorizing human-technology relations only in terms of technologies external to the body. Cyborg theory is problematic as well because most studies tend to conceptualize the cyborg merely as a discursive entity and silence the voices of people living as cyborgs. Inspired by feminist research that foregrounds the materiality of the lived and intimate relations between bodies and technologies, I argue that creating these intimate relations requires patients' active involvement in sustaining their hybrid bodies. Based on observations of these monitoring practices in a Dutch hospital and interviews with patients and technicians, the article shows that heart cyborgs are far from passive. On the contrary, their unique experience in sensing the entangled agencies of technologies and their own heart plays a crucial role in sustaining their hybrid bodies.

  8. Attrition and Adherence in a Web-Based Distress Management Program for Implantable Cardioverter Defibrillator Patients (WEBCARE): Randomized Controlled Trial

    DEFF Research Database (Denmark)

    Habibovic, M.; Cuijpers, P.; Alings, M.

    2014-01-01

    Background: WEB-Based Distress Management Program for Implantable CARdioverter defibrillator Patients (WEBCARE) is a Web-based randomized controlled trial, designed to improve psychological well-being in patients with an implantable cardioverter defibrillator (ICD). As in other Web-based trials, ...

  9. Effect of Metoprolol Versus Carvedilol on Outcomes in MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy)

    DEFF Research Database (Denmark)

    Ruwald, Martin H; Ruwald, Anne-Christine H; Jøns, Christian

    2013-01-01

    This study sought to compare the effects of metoprolol and carvedilol in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy) study.......This study sought to compare the effects of metoprolol and carvedilol in the MADIT-CRT (Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy) study....

  10. Virtual automation.

    Science.gov (United States)

    Casis, E; Garrido, A; Uranga, B; Vives, A; Zufiaurre, C

    2001-01-01

    Total laboratory automation (TLA) can be substituted in mid-size laboratories by a computer sample workflow control (virtual automation). Such a solution has been implemented in our laboratory using PSM, software developed in cooperation with Roche Diagnostics (Barcelona, Spain), to this purpose. This software is connected to the online analyzers and to the laboratory information system and is able to control and direct the samples working as an intermediate station. The only difference with TLA is the replacement of transport belts by personnel of the laboratory. The implementation of this virtual automation system has allowed us the achievement of the main advantages of TLA: workload increase (64%) with reduction in the cost per test (43%), significant reduction in the number of biochemistry primary tubes (from 8 to 2), less aliquoting (from 600 to 100 samples/day), automation of functional testing, drastic reduction of preanalytical errors (from 11.7 to 0.4% of the tubes) and better total response time for both inpatients (from up to 48 hours to up to 4 hours) and outpatients (from up to 10 days to up to 48 hours). As an additional advantage, virtual automation could be implemented without hardware investment and significant headcount reduction (15% in our lab).

  11. Antitachycardia pacing programming in implantable cardioverter defibrillator: A systematic review.

    Science.gov (United States)

    De Maria, Elia; Giacopelli, Daniele; Borghi, Ambra; Modonesi, Letizia; Cappelli, Stefano

    2017-05-26

    Implantable cardioverter defibrillator (ICD) programming involves several parameters. In recent years antitachycardia pacing (ATP) has gained an increasing importance in the treatment of ventricular arrhythmias, whether slow or fast. It reduces the number of unnecessary and inappropriate shocks and improves both patient's quality of life and device longevity. There is no clear indication regarding the type of ATP to be used, except for the treatment of fast ventricular tachycardias (188 bpm-250 bpm) where it has been shown a greater efficacy and safety of burst compared to ramp; 8 impulses in each sequence of ATP appears to be the best programming option in this setting. Beyond ATP use, excellent clinical results were obtained with programming standardization following these principles: extended detection time in ventricular fibrillation (VF) zone; supraventricular discrimination criteria up to 200 bpm; first shock in VF zone at the maximum energy in order to reduce the risk of multiple shocks. The MADIT-RIT trial and some observational registries have also recently demonstrated that programming with a widespread use of ATP, higher cut-off rates or delayed intervention reduces the number of inappropriate and unnecessary therapies and improves the survival of patients during mid-term follow-up.

  12. Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillators.

    Science.gov (United States)

    Alam, Mian Bilal; Munir, Muhammad Bilal; Rattan, Rohit; Flanigan, Susan; Adelstein, Evan; Jain, Sandeep; Saba, Samir

    2014-02-01

    Cardiac resynchronization therapy (CRT) implantable cardioverter defibrillators (ICDs) deliver high burden ventricular pacing to heart failure patients, which has a significant effect on battery longevity. The aim of this study was to investigate whether battery longevity is comparable for CRT-ICDs from different manufacturers in a contemporary cohort of patients. All the CRT-ICDs implanted at our institution from 1 January 2008 to 31 December 2010 were included in this analysis. Baseline demographic and clinical data were collected on all patients using the electronic medical record. Detailed device information was collected on all patients from scanned device printouts obtained during routine follow-up. The primary endpoint was device replacement for battery reaching the elective replacement indicator (ERI). A total of 646 patients (age 69 ± 13 years), implanted with CRT-ICDs (Boston Scientific 173, Medtronic 416, and St Jude Medical 57) were included in this analysis. During 2.7 ± 1.5 years follow-up, 113 (17%) devices had reached ERI (Boston scientific 4%, Medtronic 25%, and St Jude Medical 7%, P battery was significantly worse for Medtronic devices compared with devices from other manufacturers (94% for Boston scientific, 67% for Medtronic, and 92% for St Jude Medical, P battery longevity by manufacturer was independent of pacing burden, lead parameters, and burden of ICD therapy. There are significant discrepancies in CRT-ICD battery longevity by manufacturer. These data have important implications on clinical practice and patient outcomes.

  13. Y2K: effects on pacemaker and implantable defibrillator programmers.

    Science.gov (United States)

    Flynn, D P; Daubert, J P; Huang, D T; Ocampo, C M; O'Gorman, E

    1999-01-01

    All permanent pacemakers and implantable defibrillators (PPM/ICDs) will continue to function as programmed without regard to the date in the year 2000 (Y2K). All manufacturers contacted reassured us that some of these devices incorporate a day/year clock in the circuitry; however, these are not involved in sensing or delivering programmed therapy. Some manufacturers' device programmers will roll over to the year 2000 without any problems at all, whereas others may have difficulty with date and time stamping on printed reports. We tested 14 different types of PPM/ICD programmers for Y2K compliance using 8 tests. Five of the 14 models passed each test and were labeled at our institution with a green "Y2K" sticker to identify them as Y2K compatible and needing no special attention after December 31, 1999. The most common test failed was the ability to roll the date forward from December 31, 1999, with the programmer power off. Organizations should consider testing and replacing noncompliant device programmers or placing a red sticker with "Y2K" crossed out on noncompliant pieces. The red sticker alerts the advanced practice nurse or physician to the need to confirm the appropriate date and time in the programmer after startup in the year 2000 and before interrogating or programming any PPM/ICD, to avoid inappropriate date and time stamping on printed reports from that programmer.

  14. Implanted Cardiac Defibrillator Care in Radiation Oncology Patient Population

    International Nuclear Information System (INIS)

    Gelblum, Daphna Y.; Amols, Howard

    2009-01-01

    Purpose: To review the experience of a large cancer center with radiotherapy (RT) patients bearing implantable cardiac defibrillators (ICDs) to propose some preliminary care guidelines as we learn more about the devices and their interaction with the therapeutic radiation environment. Methods and Materials: We collected data on patients with implanted ICDs treated with RT during a 2.5-year period at any of the five Memorial Sloan-Kettering clinical campuses. Information regarding the model, location, and dose detected from the device, as well as the treatment fields, fraction size, and treatment energy was collected. During this time, a new management policy for these patients had been implemented requiring treatment with low-energy beams (6 MV) and close surveillance of the patients in partnership with their electrophysiologist, as they received RT. Results: During the study period, 33 patients were treated with an ICD in place. One patient experienced a default of the device to its initial factory setting that was detected by the patient hearing an auditory signal from the device. This patient had initially been treated with a 15-MV beam. After this episode, his treatment was replanned to be completed with 6-MV photons, and he experienced no further events. Conclusion: Patients with ICDs and other implanted computer-controlled devices will be encountered more frequently in the RT department, and proper management is important. We present a policy for the safe treatment of these patients in the radiation oncology environment.

  15. Atrial therapies reduce atrial arrhythmia burden in defibrillator patients.

    Science.gov (United States)

    Friedman, P A; Dijkman, B; Warman, E N; Xia, H A; Mehra, R; Stanton, M S; Hammill, S C

    2001-08-28

    Approximately 25% of patients who receive an implantable cardioverter-defibrillator (ICD) to treat ventricular tachyarrhythmias have documented atrial tachyarrhythmias before implantation. This study assessed the ability of device-based prevention and termination therapies to reduce the burden of spontaneous atrial tachyarrhythmias. Patients with a standard indication for the implantation of an ICD and 2 episodes of atrial tachyarrhythmias in the preceding year received a dual-chamber ICD (Medtronic 7250 Jewel AF) that uses pacing and shock therapies for prevention and/or termination of atrial tachyarrhythmias. In a multicenter trial, patients were randomized to 3-month periods with atrial therapies "on" or "off" and subsequently crossed over. Analysis was performed on the 52 of 269 patients who had episodes of atrial tachyarrhythmia and had >/=30 days of follow-up with atrial therapies on and off. The atrial therapies resulted in a reduction of atrial tachyarrhythmia burden from a mean of 58.5 to 7.8 h/mo. A paired analysis (Wilcoxon signed-rank test) showed that the median difference in burden (1.1 h/mo) was highly significant (P=0.007). When the subgroup of 41 patients treated only with atrial pacing therapies was analyzed, the reduction in burden persisted (P=0.01). In this study, patients with a standard ICD indication and atrial tachyarrhythmias had a significant reduction in atrial tachyarrhythmia burden with use of atrial pacing and shock therapies.

  16. Basic study on a lower-energy defibrillation method using computer simulation and cultured myocardial cell models.

    Science.gov (United States)

    Yaguchi, A; Nagase, K; Ishikawa, M; Iwasaka, T; Odagaki, M; Hosaka, H

    2006-01-01

    Computer simulation and myocardial cell models were used to evaluate a low-energy defibrillation technique. A generated spiral wave, considered to be a mechanism of fibrillation, and fibrillation were investigated using two myocardial sheet models: a two-dimensional computer simulation model and a two-dimensional experimental model. A new defibrillation technique that has few side effects, which are induced by the current passing into the patient's body, on cardiac muscle is desired. The purpose of the present study is to conduct a basic investigation into an efficient defibrillation method. In order to evaluate the defibrillation method, the propagation of excitation in the myocardial sheet is measured during the normal state and during fibrillation, respectively. The advantages of the low-energy defibrillation technique are then discussed based on the stimulation timing.

  17. Spinal cord stimulation for refractory angina in patients implanted with cardioverter defibrillators: five case reports

    DEFF Research Database (Denmark)

    Enggaard, Thomas P; Andersen, Claus; Scherer, Christian

    2010-01-01

    Patients implanted with a cardioverter defibrillator (ICD) who are suffering from refractory angina pectoris could benefit from spinal cord stimulation (SCS) due to the well-documented pain relieving effect. However, the combined treatment remains controversial. The aim of the study is to report...... successful long-term treatment with SCS in five patients implanted with cardioverter defibrillators. The combined treatments with ICD and thoracic epidural electrical stimulation were used in five patients with refractory angina pectoris. During the procedure of the implantation, testing with the maximal...... for refractory angina pectoris can be performed in patients implanted with cardioverter defibrillators without interference. However, individual testing during implantation or re-programming the devices is mandatory in order to assess optimal safety in each patient....

  18. Automating Finance

    Science.gov (United States)

    Moore, John

    2007-01-01

    In past years, higher education's financial management side has been riddled with manual processes and aging mainframe applications. This article discusses schools which had taken advantage of an array of technologies that automate billing, payment processing, and refund processing in the case of overpayment. The investments are well worth it:…

  19. Library Automation.

    Science.gov (United States)

    Husby, Ole

    1990-01-01

    The challenges and potential benefits of automating university libraries are reviewed, with special attention given to cooperative systems. Aspects discussed include database size, the role of the university computer center, storage modes, multi-institutional systems, resource sharing, cooperative system management, networking, and intelligent…

  20. Current state of knowledge and experts' perspective on the subcutaneous implantable cardioverter-defibrillator.

    Science.gov (United States)

    Santini, Massimo; Cappato, Riccardo; Andresen, Dietrich; Brachmann, Johannes; Davies, D Wyn; Cleland, John; Filippi, Alessandro; Gronda, Edoardo; Hauer, Richard; Steinbeck, Gerhard; Steinhaus, David

    2009-06-01

    ICD implantation is today a well-recognized therapy to prevent sudden cardiac death. The available implantable devices at present need the use of permanent endocavitary leads which may cause, in some instances, serious troubles to the patients (lead dislodgement, ventricular perforation, lead infections, etc.). A new implantable defibrillator provided by only a subcutaneous lead is at present under evaluation. Its potential indications, usefulness benefits, and problems represent an interesting field of investigation and discussion. This paper describes the conclusions recently reached by a panel of experts, with regard to the potential role of an implantable subcutaneous defibrillator in the prevention of sudden cardiac death.

  1. Defibrillator patients should not be denied a peaceful death.

    Science.gov (United States)

    Westerdahl, Annika Kinch; Sutton, Richard; Frykman, Viveka

    2015-03-01

    Implantable defibrillators (ICDs) prevent sudden cardiac death. With declining health, ICD therapy may prolong death and expose the patient to unnecessary pain and anxiety. Few studies have addressed end of life care in ICD patients. The objective of this study was to investigate end of life in ICD patients, with respect to location of death; duration between do-not-resuscitate (DNR)-orders and deactivation of ICD therapy or DNR and time of death. A descriptive analysis of 65 deceased ICD patients, all whom had a written DNR-order before death, is presented. The majority (86%) was treated in hospitals, mainly (63%) university hospitals, and many (33%) in cardiology wards. Despite DNR-order, ICD shock therapy was active in 51% of all patients. In those with therapy deactivated at death, therapy deactivation was carried out two days or more after DNR-order in more than a third (38%). The time from DNR decision to death in patients with therapy active had a median of four days (IQR 1-38). During the last 24h of life, 24% of the patients experienced shock treatment. The majority of ICD patients with a DNR-order were treated in university hospitals. More than half still had shock treatment active at time of death with a median of four days or more between DNR decision and death. Patients with therapy deactivated, two days or more elapsed in more than a third from DNR decision to deactivation of therapy, exposing patients to a high risk of painful shocks before death. Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

  2. Dual chamber arrhythmia detection in the implantable cardioverter defibrillator.

    Science.gov (United States)

    Dijkman, B; Wellens, H J

    2000-10-01

    Dual chamber implantable cardioverter defibrillator (ICD) technology extended ICD therapy to more than termination of hemodynamically unstable ventricular tachyarrhythmias. It created the basis for dual chamber arrhythmia management in which dependable detection is important for treatment and prevention of both ventricular and atrial arrhythmias. Dual chamber detection algorithms were investigated in two Medtronic dual chamber ICDs: the 7250 Jewel AF (33 patients) and the 7271 Gem DR (31 patients). Both ICDs use the same PR Logic algorithm to interpret tachycardia as ventricular tachycardia (VT), supraventricular tachycardia (SVT), or dual (VT+ SVT). The accuracy of dual chamber detection was studied in 310 of 1,367 spontaneously occurring tachycardias in which rate criterion only was not sufficient for arrhythmia diagnosis. In 78 episodes there was a double tachycardia, in 223 episodes SVT was detected in the VT or ventricular fibrillation zone, and in 9 episodes arrhythmia was detected outside the boundaries of the PR Logic functioning. In 100% of double tachycardias the VT was correctly diagnosed and received priority treatment. SVT was seen in 59 (19%) episodes diagnosed as VT. The causes of inappropriate detection were (1) algorithm failure (inability to fulfill the PR

  3. Comparison of Sprint Fidelis and Riata defibrillator lead failure rates.

    Science.gov (United States)

    Fazal, Iftikhar A; Shepherd, Ewen J; Tynan, Margaret; Plummer, Christopher J; McComb, Janet M

    2013-09-30

    Sprint Fidelis and Riata defibrillator leads are prone to early failure. Few data exist on the comparative failure rates and mortality related to lead failure. The aims of this study were to determine the failure rate of Sprint Fidelis and Riata leads, and to compare failure rates and mortality rates in both groups. Patients implanted with Sprint Fidelis leads and Riata leads at a single centre were identified and in July 2012, records were reviewed to ascertain lead failures, deaths, and relationship to device/lead problems. 113 patients had Sprint Fidelis leads implanted between June 2005 and September 2007; Riata leads were implanted in 106 patients between January 2003 and February 2008. During 53.0 ± 22.3 months of follow-up there were 13 Sprint Fidelis lead failures (11.5%, 2.60% per year) and 25 deaths. Mean time to failure was 45.1 ± 15.5 months. In the Riata lead cohort there were 32 deaths, and 13 lead failures (11.3%, 2.71% per year) over 54.8 ± 26.3 months follow-up with a mean time to failure of 53.5 ± 24.5 months. There were no significant differences in the lead failure-free Kaplan-Meier survival curve (p=0.77), deaths overall (p=0.17), or deaths categorised as sudden/cause unknown (p=0.54). Sprint Fidelis and Riata leads have a significant but comparable failure rate at 2.60% per year and 2.71% per year of follow-up respectively. The number of deaths in both groups is similar and no deaths have been identified as being related to lead failure in either cohort. Copyright © 2012. Published by Elsevier Ireland Ltd.

  4. Cardiogenic Shock and Lung Injury as a Complication of Defibrillation

    Directory of Open Access Journals (Sweden)

    Hasan Serdar Kıhtır

    2017-12-01

    Full Text Available Local burns, embolism, and arrhythmia are the most common side effects observed after electrical shock treatments. However, systolic function may be rarely affected and pulmonary edema may develop. The cases of pulmonary edema after electrical shock treatments have been reported since 1960s and the proposed mechanism is the inadequacy of the left atrium cuff and ventricle. It was learned that a 7-year-old-girl without any known disease except vesicoureteral reflux had a ventricular fibrillation during general anesthesia induction and defibrillation at 2 joule/kg was attempted. It was also learned that the procedure was delayed and the patient was diagnosed with a long QT (QTc: 0.47 ms and had respiratory distress and circulatory disturbances after four hours. Pulmonary edema and heart failure was determined, and due to hipoxemia (SpO2 <88% not getting any better with non-invasive ventilation, the patient was intubated and followed with mechanical ventilation. A thermodilution catheter was inserted into the femoral artery and a low cardiac index (CI: 1.58 L/min/m2, elevated extravascular lung water index (EVLWI: 18 mL/kg and high pulmonary vascular permeability index (PVPI: 7.6 were determined. The patient was treated by mechanical ventilation and vasoactive/inotropic management and discharged at the fifth day of hospitalization without any sequela. Having high EVLWI with high PVPI suggest that the pulmonary edema mechanism may also be caused by alveolocapillary membrane damage, which is not accompanied by heart failure alone. This case is presented to show that it is the first child in the literature and that the results of transpulmonary thermodilution can also give information about lung function as well as cardiac function.

  5. The effect of intermittent atrial tachyarrhythmia on heart failure or death in cardiac resynchronization therapy with defibrillator versus implantable cardioverter-defibrillator patients

    DEFF Research Database (Denmark)

    Ruwald, Anne-Christine; Pietrasik, Grzegorz; Goldenberg, Ilan

    2014-01-01

    OBJECTIVES: This study aimed to investigate the effect of both history of intermittent atrial tachyarrhythmias (IAT) and in-trial IAT on the risk of heart failure (HF) or death comparing cardiac resynchronization therapy with defibrillator (CRT-D) to implantable cardioverter-defibrillator (ICD......) treatment in mildly symptomatic HF patients with left bundle branch block (LBBB). BACKGROUND: Limited data exist regarding the benefit of CRT-D in patients with IAT. METHODS: The benefit of CRT-D in reducing the risk of HF/death was evaluated using multivariate Cox models incorporating the presence of......-D versus ICD on the risk of HF/death was not significantly different between LBBB patients with or without history of IAT (HR: 0.50, p = 0.028, and HR: 0.46, p

  6. Wearable cardioverter defibrillators for the prevention of sudden cardiac arrest: a health technology assessment and patient focus group study

    Directory of Open Access Journals (Sweden)

    Ettinger S

    2017-11-01

    -to-medium term, but the quality of existing evidence is very low. AEs and SAEs need to be more appropriately reported in order to further evaluate the safety of the device. High-quality comparative evidence and well-described disease groups are required to assess the effectiveness of the WCD and to determine which patient groups may benefit most from the intervention. Keywords: sudden cardiac arrest, ventricular tachycardia, ventricular fibrillation, cardioverter defibrillator, external, wearable, patient involvement

  7. Monitoring device acceptance in implantable cardioverter defibrillator patients using the Florida Patient Acceptance Survey

    DEFF Research Database (Denmark)

    Versteeg, Henneke; Starrenburg, Annemieke; Denollet, Johan

    2012-01-01

    Patient device acceptance might be essential in identifying patients at risk for adverse patient-reported outcomes following implantation of an implantable cardioverter defibrillator (ICD). We examined the validity and reliability of the Florida Patient Acceptance Scale (FPAS) and identified corr...

  8. Gender disparities in anxiety and quality of life in patients with an implantable cardioverter-defibrillator

    DEFF Research Database (Denmark)

    Habibović, Mirela; van den Broek, Krista C; Theuns, Dominic A M J

    2011-01-01

    A paucity of studies in implantable cardioverter-defibrillator (ICD) patients has examined gender disparities in patient-reported outcomes, such as anxiety and quality of life (QoL). We investigated (i) gender disparities in anxiety and QoL and (ii) the magnitude of the effect of gender vs. New...

  9. E-Health to Manage Distress in Patients With an Implantable Cardioverter-Defibrillator

    DEFF Research Database (Denmark)

    Habibović, Mirela; Denollet, Johan; Cuijpers, Pim

    2014-01-01

    UNLABELLED: The Web-based distress management program for patients with an implantable cardioverter-defibrillator (ICD; WEBCARE) was developed to mitigate distress and enhance health-related quality of life in ICD patients. This study investigated the treatment effectiveness at 3-month follow-up ...

  10. Implantable cardioverter defibrillator therapy for prevention of sudden cardiac death in children in the Netherlands

    NARCIS (Netherlands)

    Heersche, Jogien H. M.; Blom, Nico A.; van de Heuvel, Freek; Blank, Christiaan; Reimer, Annette G.; Clur, Sally-Ann; Witsenburg, Maarten; ten Harkel, A. Derk Jan

    2010-01-01

    INTRODUCTION: Implantable cardioverter defibrillator (ICD) therapy is increasingly used in children. The purpose of this multicenter study is to evaluate mid-term clinical outcome and to identify predictors for device discharge in pediatric ICD recipients. METHODS AND RESULTS: From 1995 to 2006, 45

  11. SUDDEN CARDIAC DEATH AND THE USE OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATORS IN PEDIATRIC-PATIENTS

    NARCIS (Netherlands)

    SILKA, MJ; KRON, J; DUNNIGAN, A; DICK, M; BINKBOELKENS, M; ERICKSON, CC; JEDEIKIN, R; WETZEL, GT; VANHARE, GF; CAMPBELL, R; WALSH, E; SAUL, JP; SCHAFFER, MS; KARPAWICH, P; VOGEL, RL; BENSON, DW; DEAL, B; SCAGLIOTTI, D; STERBA, R; HORDOF, AJ; KRONGRAD, E; KANTER, RJ; EPSTEIN, M; COHEN, M; BEDER, S; HAMILTON, R; FOURNIER, A; HUBBARD, J; CHRISTIANSEN, JL; JENNINGS, J; VILLAFANE, J; PORTER, CBJ; CASE, C; GILLETTE, PC; BELAND, M; KUGLER, JD; OCONNOR, BK; ALLENDER, H; HERNDON, SP; SMITH, RT; BURTON, D; KURER, CC; BYRUM, C; GUAM, WE; FRIEDMAN, R; PERRY, JC; SCOTT, W; MEHTA, AV; PICKHOFF, AS; FISH, F; YEAGER, S; KAWABORI, [No Value; TRIPPLE, M; ROSENFELD, LE

    Background. During the past decade. the implantable cardioverter-defibrillator (ICD) has emerged as the primary therapeutic option for survivors of sudden cardiac death (SCD). Investigation of the clinical efficacy of these devices has primarily assessed outcome in adults with coronary artery

  12. Implantable Cardioverter Defibrillator Therapy for Prevention of Sudden Cardiac Death in Children in The Netherlands

    NARCIS (Netherlands)

    Heersche, Jogien H. M.; Blom, Nico A.; Van De Heuvel, Freek; Blank, Christiaan; Reimer, Annette G.; Clur, Sally-Ann; Witsenburg, Maarten; Ten Harkel, A. Derk Jan

    Introduction: Implantable cardioverter defibrillator (ICD) therapy is increasingly used in children. The purpose of this multicenter study is to evaluate mid-term clinical outcome and to identify predictors for device discharge in pediatric ICD recipients. Methods and Results: From 1995 to 2006, 45

  13. End-tidal carbon dioxide and defibrillation success in out-of-hospital cardiac arrest.

    Science.gov (United States)

    Savastano, Simone; Baldi, Enrico; Raimondi, Maurizio; Palo, Alessandra; Belliato, Mirko; Cacciatore, Elisa; Corazza, Valentina; Molinari, Simone; Canevari, Fabrizio; Danza, Aurora I; De Ferrari, Gaetano M; Iotti, Giorgio Antonio; Visconti, Luigi Oltrona

    2017-12-01

    Basing on the relationship between the quality of cardiopulmonary resuscitation (CPR) and the responsiveness of VF to the defibrillation we aimed to assess whether the values of ETCO2 in the minute before defibrillation could predict the effectiveness of the shock. We retrospectively evaluated the reports generated by the manual monitor/defibrillator (Corpuls by GS Elektromedizinische Geräte G. Stemple GmbH, Germany) used for cases of VF cardiac arrest from January 2015 to December 2016. The mean ETCO2 value of the minute preceding the shock (METCO2 60 ) was computed. A blind evaluation of the effectiveness of each shock was provided by three cardiologists. A total amount of 207 shocks were delivered for 62 patients. When considering the three tertiles of METCO2 60 (T1:METCO2 60 ≤ 20mmHg; T2: 20mmHg 31mmHg) a statistically significant difference between the percentages of shock success was found (T1: 50%; T2: 63%; T3: 78%; Chi square p=0.003; p for trend CPR, monitored via ETCO2, and suggest ETCO2 monitoring as an additional weapon to guide defibrillation. Copyright © 2017 Elsevier B.V. All rights reserved.

  14. Undertreatment of anxiety and depression in patients with an implantable cardioverter-defibrillator

    DEFF Research Database (Denmark)

    Hoogwegt, Madelein T; Kupper, Nina; Theuns, Dominic A M J

    2012-01-01

    Twenty-five to 33% of patients with an implantable cardioverter-defibrillator (ICD) experience anxiety and depression, but it is not known whether their symptoms are adequately treated. We investigated (a) whether patients with clinically relevant symptoms of distress received appropriate treatme......, and (b) whether patients not treated for their emotional distress reported poorer health status using a prospective study design....

  15. Experiences of patients with implantable cardioverter defibrillator in Turkey: A qualitative study

    Directory of Open Access Journals (Sweden)

    Özgür Aslan

    2012-01-01

    Full Text Available Background. There has been an increase in the number of patients with implantable cardioverter defibrillator (ICD. It isimportant to understand ICD patients’ experiences with it.Aim. The aim of this study was to describe experiences of patients with implantable cardioverter defibrillator (ICD.Methodology. A qualitative approach was used. Focus group interviews were used to obtain data from 19 patients whowere implanted cardioverter defibrillator at two centers in Izmir, Turkey. The patients were assigned into four groups. Thedata was analyzed using qualitative content analysis.Results. The analysis revealed six main themes: activities of daily living, social life, familial relationships, emotionalchanges, implantable cardioverter defibrillator shocks and experiences with receiving information and counselling fromhealth care providers.Conclusions. It can be concluded that patients with ICD experience physical and psychological problems and are not offeredthe education they need. To reduce the fears of the patients and their families and to prepare them for possible life stylechanges, comprehensive training programs that start in the pre-implantation period and continue into the post-implantationperiod should be organized.

  16. Increased anxiety in partners of patients with a cardioverter-defibrillator

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; VAN DEN Berg, Martha; Erdman, Ruud A M

    2009-01-01

    The partner of the implantable cardioverter-defibrillator (ICD) patient serves as an important source of support for the patient, which may be hampered if the partner experiences increased distress. We examined (1) potential differences in anxiety and depressive symptoms in ICD patients compared...

  17. The learning curve associated with the introduction of the subcutaneous implantable defibrillator

    NARCIS (Netherlands)

    Knops, Reinoud E.; Brouwer, Tom F.; Barr, Craig S.; Theuns, Dominic A.; Boersma, Lucas; Weiss, Raul; Neuzil, Petr; Scholten, Marcoen; Lambiase, Pier D.; Leon, Angel R.; Hood, Margaret; Jones, Paul W.; Wold, Nicholas; Grace, Andrew A.; Olde Nordkamp, Louise R. A.; Burke, Martin C.

    2016-01-01

    Aims The subcutaneous implantable cardioverter defibrillator (S-ICD) was introduced to overcome complications related to transvenous leads. Adoption of the S-ICD requires implanters to learn a new implantation technique. The aim of this study was to assess the learning curve for S-ICD implanters

  18. The learning curve associated with the introduction of the subcutaneous implantable defibrillator

    NARCIS (Netherlands)

    R.E. Knops (Reinoud); T.F. Brouwer (Tom F.); C.S. Barr (Craig); D.A.M.J. Theuns (Dominic); L. Boersma (Lucas); R. Weiss (Ram); P. Neuzil (Petr); M.F. Scholten (Marcoen); P.D. Lambiase (Pier); A. Leon (Angel); A.M. Hood (Margaret); P. Jones; Wold, N. (Nicholas); Grace, A.A. (Andrew A.); L.R.A. Olde Nordkamp (Louise R.A.); M.C. Burke (Martin)

    2016-01-01

    textabstractAims: The subcutaneous implantable cardioverter defibrillator (S-ICD) was introduced to overcome complications related to transvenous leads. Adoption of the S-ICD requires implanters to learn a new implantation technique. The aim of this study was to assess the learning curve for S-ICD

  19. Long-Term Clinical Outcomes of Subcutaneous Versus Transvenous Implantable Defibrillator Therapy

    NARCIS (Netherlands)

    Brouwer, Tom F.; Yilmaz, Dilek; Lindeboom, Robert; Buiten, Maurits S.; Olde Nordkamp, Louise R. A.; Schalij, Martin J.; Wilde, Arthur A.; van Erven, Lieselot; Knops, Reinoud E.

    2016-01-01

    Transvenous implantable cardioverter-defibrillators (TV-ICDs) improve survival in patients at risk for sudden cardiac death, but complications remain an important drawback. The subcutaneous ICD (S-ICD) was developed to overcome lead-related complications. Comparison of clinical outcomes of both

  20. Anxiety and risk of ventricular arrhythmias or mortality in patients with an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Habibović, Mirela; Pedersen, Susanne S.; van den Broek, Krista C

    2013-01-01

    A subgroup of patients with an implantable cardioverter defibrillator (ICD) experiences anxiety after device implantation. The purpose of the present study was to evaluate whether anxiety is predictive of ventricular arrhythmias and all-cause mortality 1 year post ICD implantation....

  1. Ventricular Tachyarrhythmias and Mortality in Patients With an Implantable Cardioverter Defibrillator

    DEFF Research Database (Denmark)

    Mastenbroek, Mirjam H; Versteeg, Henneke; Jordaens, Luc

    2014-01-01

    Objective We examined whether depression is independently associated with implantable cardioverter defibrillator (ICD) therapy for ventricular tachyarrhythmias and mortality. Methods A cohort of 430 consecutive patients with a first-time ICD (79% men; mean [standard deviation] age = 57.8 [12.1] y...

  2. Inappropriate shock delivery by implantable cardioverter defibrillator due to electrical interference with washing machine.

    Science.gov (United States)

    Chongtham, Dhanaraj Singh; Bahl, Ajay; Kumar, Rohit Manoj; Talwar, K K

    2007-05-31

    We report a patient with hypertrophic cardiomyopathy who received an inappropriate implantable cardioverter defibrillator shock due to electrical interference from a washing machine. This electrical interference was detected as an episode of ventricular fibrillation with delivery of shock without warning symptoms.

  3. Success and failure of the defibrillation shock: insights from a simulation study.

    Science.gov (United States)

    Skouibine, K; Trayanova, N; Moore, P

    2000-07-01

    This simulation study presents a further inquiry into the mechanisms by which a strong electric shock fails to halt life-threatening cardiac arrhythmias. The research uses a model of the defibrillation process that represents a sheet of myocardium as a bidomain. The tissue consists of nonuniformly curved fibers in which spiral wave reentry is initiated. Monophasic defibrillation shocks are delivered via two line electrodes that occupy opposite tissue boundaries. In some simulation experiments, the polarity of the shock is reversed. Electrical activity in the sheet is compared for failed and successful shocks under controlled conditions. The maps of transmembrane potential and activation times calculated during and after the shock demonstrate that weak shocks fail to terminate the reentrant activity via two major mechanisms. As compared with strong shocks, weak shocks result in (1) smaller extension of refractoriness in the areas depolarized by the shock, and (2) slower or incomplete activation of the excitable gap created by deexcitation of the negatively polarized areas. In its turn, mechanism 2 is associated with one or more of the following events: (a) lack of some break excitations, (b) latency in the occurrence of the break excitations, and (c) slower propagation through deexcited areas. Reversal of shock polarity results in a change of the extent of the regions of deexcitation, and thus, in a change in defibrillation threshold. The results of this study indicate the paramount importance of shock-induced deexcitation in both defibrillation and postshock arrhythmogenesis.

  4. Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance

    NARCIS (Netherlands)

    Tjong, F. V. Y.; Brouwer, T. F.; Smeding, L.; Kooiman, K. M.; de Groot, J. R.; Ligon, D.; Sanghera, R.; Schalij, M. J.; Wilde, A. A. M.; Knops, R. E.

    2016-01-01

    The subcutaneous implantable cardioverter-defibrillator (S-ICD) and leadless pacemaker (LP) are evolving technologies that do not require intracardiac leads. However, interactions between these two devices are unexplored. We investigated the feasibility, safety, and performance of combined LP and

  5. Cardiac pacemaker battery discharge after external electrical cardioversion for broad QRS Complex Tachycardia.

    Science.gov (United States)

    Annamaria, Martino; Andrea, Scapigliati; Michela, Casella; Tommaso, Sanna; Gemma, Pelargonio; Antonio, Dello Russo; Roberto, Zamparelli; Stefano, De Paulis; Fulvio, Bellocci; Rocco, Schiavello

    2008-08-01

    External electrical cardioversion or defibrillation may be necessary in patients with implanted cardiac pacemaker (PM) or implantable cardioverter defibrillator (ICD). Sudden discharge of high electrical energy employed in direct current (DC) transthoracic countershock may damage the PM/ICD system resulting in a series of possible device malfunctions. For this reason, when defibrillation or cardioversion must be attempted in a patient with a PM or ICD, some precautions should be taken, particularly in PM dependent patients, in order to prevent damage to the device. We report the case of a 76-year-old woman with a dual chamber PM implanted in the right subclavicular region, who received two consecutive transthoracic DC shocks to treat haemodynamically unstable broad QRS complex tachycardia after cardiac surgery performed with a standard sternotomic approach. Because of the sternal wound and thoracic drainage tubes together with the severe clinical compromise, the anterior paddle was positioned near the pulse generator. At the following PM test, a complete battery discharge was detected.

  6. Implantable Cardioverter-defibrillator Therapy for Hypertrophic Cardiomyopathy: Usefulness in Primary and Secondary Prevention.

    Science.gov (United States)

    Sarrias, Axel; Galve, Enrique; Sabaté, Xavier; Moya, Àngel; Anguera, Ignacio; Núñez, Elaine; Villuendas, Roger; Alcalde, Óscar; García-Dorado, David

    2015-06-01

    Hypertrophic cardiomyopathy is a frequent cause of sudden death. Clinical practice guidelines indicate defibrillator implantation for primary prevention in patients with 1 or more risk factors and for secondary prevention in patients with a history of aborted sudden death or sustained ventricular arrhythmias. The aim of the present study was to analyze the follow-up of patients who received an implantable defibrillator following the current guidelines in nonreferral centers for this disease. This retrospective observational study included all patients who underwent defibrillator implantation between January 1996 and December 2012 in 3 centers in the province of Barcelona. The study included 69 patients (mean age [standard deviation], 44.8 [17] years; 79.3% men), 48 in primary prevention and 21 in secondary prevention. The mean number of risk factors per patient was 1.8 in the primary prevention group and 0.5 in the secondary prevention group (P=.029). The median follow-up duration was 40.5 months. The appropriate therapy rate was 32.7/100 patient-years in secondary prevention and 1.7/100 patient-years in primary prevention (P<.001). Overall mortality was 10.1%. Implant-related complications were experienced by 8.7% of patients, and 13% had inappropriate defibrillator discharges. In patients with a defibrillator for primary prevention, the appropriate therapy rate is extremely low, indicating the low predictive power of the current risk stratification criteria. Copyright © 2014 Sociedad Española de Cardiología. Published by Elsevier España, S.L.U. All rights reserved.

  7. "Real life" longevity of implantable cardioverter-defibrillator devices.

    Science.gov (United States)

    Manolis, Antonis S; Maounis, Themistoklis; Koulouris, Spyridon; Vassilikos, Vassilios

    2017-09-01

    Manufacturers of implantable cardioverter-defibrillators (ICDs) promise a 5- to 9-year projected longevity; however, real-life data indicate otherwise. The aim of the present study was to assess ICD longevity among 685 consecutive patients over the last 20 years. Real-life longevity of ICDs may differ from that stated by the manufacturers. The study included 601 men and 84 women (mean age, 63.1 ± 13.3 years). The underlying disease was coronary (n = 396) or valvular (n = 15) disease, cardiomyopathy (n = 220), or electrical disease (n = 54). The mean ejection fraction was 35%. Devices were implanted for secondary (n = 562) or primary (n = 123) prevention. Single- (n = 292) or dual-chamber (n = 269) or cardiac resynchronization therapy (CRT) devices (n = 124) were implanted in the abdomen (n = 17) or chest (n = 668). Over 20 years, ICD pulse generator replacements were performed in 238 patients (209 men; age 63.7 ± 13.9 years; ejection fraction, 37.7% ± 14.0%) who had an ICD for secondary (n = 210) or primary (n = 28) prevention. The mean ICD longevity was 58.3 ± 18.7 months. In 20 (8.4%) patients, devices exhibited premature battery depletion within 36 months. Most (94%) patients had none, minor, or modest use of ICD therapy. Longevity was longest for single-chamber devices and shortest for CRT devices. Latest-generation devices replaced over the second decade lasted longer compared with devices replaced during the first decade. When analyzed by manufacturer, Medtronic devices appeared to have longer longevity by 13 to 18 months. ICDs continue to have limited longevity of 4.9 ± 1.6 years, and 8% demonstrate premature battery depletion by 3 years. CRT devices have the shortest longevity (mean, 3.8 years) by 13 to 17 months, compared with other ICD devices. These findings have important implications, particularly in view of the high expense involved with this type of electrical

  8. Plant automation

    International Nuclear Information System (INIS)

    Christensen, L.J.; Sackett, J.I.; Dayal, Y.; Wagner, W.K.

    1989-01-01

    This paper describes work at EBR-II in the development and demonstration of new control equipment and methods and associated schemes for plant prognosis, diagnosis, and automation. The development work has attracted the interest of other national laboratories, universities, and commercial companies. New initiatives include use of new control strategies, expert systems, advanced diagnostics, and operator displays. The unique opportunity offered by EBR-II is as a test bed where a total integrated approach to automatic reactor control can be directly tested under real power plant conditions

  9. “Atrial torsades de pointes” Induced by Low-Energy Shock From Implantable-Cardioverter Defibrillator

    Directory of Open Access Journals (Sweden)

    Ilknur Can, MD

    2013-09-01

    Full Text Available A 58 year-old-patient developed an episode of polymorphic atrial tachycardia which looked like "atrial torsades de pointes" after a 5J shock from implantable cardioverter defibrillator.

  10. Cost-Effectiveness of Adding Cardiac Resynchronization Therapy to an Implantable Cardioverter-Defibrillator Among Patients With Mild Heart Failure

    DEFF Research Database (Denmark)

    Woo, Christopher Y; Strandberg, Erika J; Schmiegelow, Michelle D

    2015-01-01

    -defibrillator (ICD) alone among patients with left ventricular systolic dysfunction, prolonged intraventricular conduction, and mild heart failure. DESIGN: Markov decision model. DATA SOURCES: Clinical trials, clinical registries, claims data from Centers for Medicare & Medicaid Services, and Centers for Disease...

  11. Late Ratchet syndrome involving isolated left ventricular lead dislodgement post-cardiac resynchronization therapy defibrillator generator change.

    Science.gov (United States)

    Tan, Vern Hsen; Wong, Kelvin

    2018-04-01

    Lead dislodgement following cardiac implantable electronic device (CIED) generator change is rare. We report a case including the postulate mechanism of an isolated left ventricular lead dislodgement 3 months after cardiac resynchronization therapy defibrillator pulse generator change.

  12. Automation from pictures

    International Nuclear Information System (INIS)

    Kozubal, A.J.

    1992-01-01

    The state transition diagram (STD) model has been helpful in the design of real time software, especially with the emergence of graphical computer aided software engineering (CASE) tools. Nevertheless, the translation of the STD to real time code has in the past been primarily a manual task. At Los Alamos we have automated this process. The designer constructs the STD using a CASE tool (Cadre Teamwork) using a special notation for events and actions. A translator converts the STD into an intermediate state notation language (SNL), and this SNL is compiled directly into C code (a state program). Execution of the state program is driven by external events, allowing multiple state programs to effectively share the resources of the host processor. Since the design and the code are tightly integrated through the CASE tool, the design and code never diverge, and we avoid design obsolescence. Furthermore, the CASE tool automates the production of formal technical documents from the graphic description encapsulated by the CASE tool. (author)

  13. Individual differences in the calibration of trust in automation.

    Science.gov (United States)

    Pop, Vlad L; Shrewsbury, Alex; Durso, Francis T

    2015-06-01

    The objective was to determine whether operators with an expectancy that automation is trustworthy are better at calibrating their trust to changes in the capabilities of automation, and if so, why. Studies suggest that individual differences in automation expectancy may be able to account for why changes in the capabilities of automation lead to a substantial change in trust for some, yet only a small change for others. In a baggage screening task, 225 participants searched for weapons in 200 X-ray images of luggage. Participants were assisted by an automated decision aid exhibiting different levels of reliability. Measures of expectancy that automation is trustworthy were used in conjunction with subjective measures of trust and perceived reliability to identify individual differences in trust calibration. Operators with high expectancy that automation is trustworthy were more sensitive to changes (both increases and decreases) in automation reliability. This difference was eliminated by manipulating the causal attribution of automation errors. Attributing the cause of automation errors to factors external to the automation fosters an understanding of tasks and situations in which automation differs in reliability and may lead to more appropriate trust. The development of interventions can lead to calibrated trust in automation. © 2014, Human Factors and Ergonomics Society.

  14. Fatigue and voluntary utilization of automation in simulated driving.

    Science.gov (United States)

    Neubauer, Catherine; Matthews, Gerald; Langheim, Lisa; Saxby, Dyani

    2012-10-01

    A driving simulator was used to assess the impact on fatigue, stress, and workload of full vehicle automation that was initiated by the driver. Previous studies have shown that mandatory use of full automation induces a state of "passive fatigue" associated with loss of alertness. By contrast, voluntary use of automation may enhance the driver's perceptions of control and ability to manage fatigue. Participants were assigned to one of two experimental conditions, automation optional (AO) and nonautomation (NA), and then performed a 35 min, monotonous simulated drive. In the last 5 min, automation was unavailable and drivers were required to respond to an emergency event. Subjective state and workload were evaluated before and after the drive. Making automation available to the driver failed to alleviate fatigue and stress states induced by driving in monotonous conditions. Drivers who were fatigued prior to the drive were more likely to choose to use automation, but automation use increased distress, especially in fatigue-prone drivers. Drivers in the AO condition were slower to initiate steering responses to the emergency event, suggesting optional automation may be distracting. Optional, driver-controlled automation appears to pose the same dangers to task engagement and alertness as externally initiated automation. Drivers of automated vehicles may be vulnerable to fatigue that persists when normal vehicle control is restored. It is important to evaluate automated systems' impact on driver fatigue, to seek design solutions to the issue of maintaining driver engagement, and to address the vulnerabilities of fatigue-prone drivers.

  15. Automated Student Aid Processing: The Challenge and Opportunity.

    Science.gov (United States)

    St. John, Edward P.

    1985-01-01

    To utilize automated technology for student aid processing, it is necessary to work with multi-institutional offices (student aid, admissions, registration, and business) and to develop automated interfaces with external processing systems at state and federal agencies and perhaps at need-analysis organizations and lenders. (MLW)

  16. Integration of Attributes from Non-Linear Characterization of Cardiovascular Time-Series for Prediction of Defibrillation Outcomes.

    Directory of Open Access Journals (Sweden)

    Sharad Shandilya

    Full Text Available The timing of defibrillation is mostly at arbitrary intervals during cardio-pulmonary resuscitation (CPR, rather than during intervals when the out-of-hospital cardiac arrest (OOH-CA patient is physiologically primed for successful countershock. Interruptions to CPR may negatively impact defibrillation success. Multiple defibrillations can be associated with decreased post-resuscitation myocardial function. We hypothesize that a more complete picture of the cardiovascular system can be gained through non-linear dynamics and integration of multiple physiologic measures from biomedical signals.Retrospective analysis of 153 anonymized OOH-CA patients who received at least one defibrillation for ventricular fibrillation (VF was undertaken. A machine learning model, termed Multiple Domain Integrative (MDI model, was developed to predict defibrillation success. We explore the rationale for non-linear dynamics and statistically validate heuristics involved in feature extraction for model development. Performance of MDI is then compared to the amplitude spectrum area (AMSA technique.358 defibrillations were evaluated (218 unsuccessful and 140 successful. Non-linear properties (Lyapunov exponent > 0 of the ECG signals indicate a chaotic nature and validate the use of novel non-linear dynamic methods for feature extraction. Classification using MDI yielded ROC-AUC of 83.2% and accuracy of 78.8%, for the model built with ECG data only. Utilizing 10-fold cross-validation, at 80% specificity level, MDI (74% sensitivity outperformed AMSA (53.6% sensitivity. At 90% specificity level, MDI had 68.4% sensitivity while AMSA had 43.3% sensitivity. Integrating available end-tidal carbon dioxide features into MDI, for the available 48 defibrillations, boosted ROC-AUC to 93.8% and accuracy to 83.3% at 80% sensitivity.At clinically relevant sensitivity thresholds, the MDI provides improved performance as compared to AMSA, yielding fewer unsuccessful defibrillations

  17. WIDAFELS flexible automation systems

    International Nuclear Information System (INIS)

    Shende, P.S.; Chander, K.P.; Ramadas, P.

    1990-01-01

    After discussing the various aspects of automation, some typical examples of various levels of automation are given. One of the examples is of automated production line for ceramic fuel pellets. (M.G.B.)

  18. An Automation Planning Primer.

    Science.gov (United States)

    Paynter, Marion

    1988-01-01

    This brief planning guide for library automation incorporates needs assessment and evaluation of options to meet those needs. A bibliography of materials on automation planning and software reviews, library software directories, and library automation journals is included. (CLB)

  19. Low cost automation

    International Nuclear Information System (INIS)

    1987-03-01

    This book indicates method of building of automation plan, design of automation facilities, automation and CHIP process like basics of cutting, NC processing machine and CHIP handling, automation unit, such as drilling unit, tapping unit, boring unit, milling unit and slide unit, application of oil pressure on characteristics and basic oil pressure circuit, application of pneumatic, automation kinds and application of process, assembly, transportation, automatic machine and factory automation.

  20. Evaluation of a novel ventricular support device with defibrillation capabilities in canine and porcine animal models.

    Science.gov (United States)

    Killingsworth, Cheryl R; Rippy, Marian K; Virmani, Renu; Rollins, Dennis L; McGiffin, David C; Ideker, Raymond E

    2008-08-01

    Sudden death is prevalent in heart failure patients. We tested an implantable ventricular support device consisting of a wireform harness with one or two pairs of integrated defibrillation electrode coils. The device was implanted into six pigs (36-44 kg) through a subxiphoid incision. Peak voltage (V) defibrillation thresholds (DFT) were determined for five test configurations compared with a control transvenous lead (RV to CanPect). Defibrillator can location (abdominal or pectoral) and common coil separation on the implant (0 degrees or 60 degrees ) were studied.(.) The DFT for RV60 to LV60 + CanPect was significantly less than control (348 +/- 57 vs 473 +/- 27 V, P < 0.05). The DFTs for other vectors were similar to control except for RV0 to LV0 + CanAbd (608 +/- 159 V). The device was implanted into 12 adult dogs for 42, 90, or 180 days with DFT and pathological examination performed at the terminal study. Cardiac pressures were determined at baseline, after implantation, and at the terminal study. The DFT was also determined in a separate group of four dogs at 42 days following implantation of the support device with one pair of defibrillation electrodes. The DFTs at implant and explant in dogs with one pair (8 +/- 1.5 Joules [J] and 6 +/- 1.9 J) or two pairs (8 +/- 3.4 J and 7 +/- 1.9 J) of defibrillation electrodes were not significantly different from each other but significantly less than control measured at the terminal study (18 +/- 3.4 J). Left-sided pressures were significantly decreased at explant but within expected normal ranges. Right-sided pressures were not different except for RV systolic. Histopathology indicated mild to moderate epicardial inflammation and fibrosis, consistent with a foreign body healing response. This defibrillation-enabled ventricular support system maintained mechanical functionality for up to 6 months while inducing typical chronic healing responses. The DFT was equal to or lower than a standard transvenous vector.

  1. Economic impact of longer battery life of cardiac resynchronization therapy defibrillators in Sweden

    Directory of Open Access Journals (Sweden)

    Gadler F

    2016-10-01

    Full Text Available Fredrik Gadler,1 Yao Ding,2 Nathalie Verin,3 Martin Bergius,4 Jeffrey D Miller,5 Gregory M Lenhart,5 Mason W Russell5 1Department of Cardiology, Karolinska University Hospital, Stockholm, Sweden; 2Truven Health Analytics, an IBM Company, Bethesda, MD, USA; 3Boston Scientific Corporation, Hemel Hempstead, Hertfordshire, UK; 4Boston Scientific Nordic AB, Helsingborg, Sweden; 5Truven Health Analytics, an IBM Company, Cambridge, MA, USA Objective: The objective of this study was to quantify the impact that longer battery life of cardiac resynchronization therapy defibrillator (CRT-D devices has on reducing the number of device replacements and associated costs of these replacements from a Swedish health care system perspective.Methods: An economic model based on real-world published data was developed to estimate cost savings and avoided device replacements for CRT-Ds with longer battery life compared with devices with industry-standard battery life expectancy. Base-case comparisons were performed among CRT-Ds of three manufacturers – Boston Scientific Corporation, St. Jude Medical, and Medtronic – over a 6-year time horizon, as per the available clinical data. As a sensitivity analysis, we evaluated CRT-Ds as well as single-chamber implantable cardioverter defibrillator (ICD-VR and dual-chamber implantable cardioverter defibrillator (ICD-DR devices over a longer 10-year period. All costs were in 2015 Swedish Krona (SEK discounted at 3% per annum.Results: Base-case analysis results show that up to 603 replacements and up to SEK 60.4 million cumulative-associated costs could be avoided over 6 years by using devices with extended ­battery life. The pattern of savings over time suggests that savings are modest initially but increase rapidly beginning in the third year of follow-up with each year’s cumulative savings two to three times the previous year. Evaluating CRT-D, ICD-VR, and ICD-DR devices together over a longer 10-year period, the

  2. Automated Budget System -

    Data.gov (United States)

    Department of Transportation — The Automated Budget System (ABS) automates management and planning of the Mike Monroney Aeronautical Center (MMAC) budget by providing enhanced capability to plan,...

  3. Automation 2017

    CERN Document Server

    Zieliński, Cezary; Kaliczyńska, Małgorzata

    2017-01-01

    This book consists of papers presented at Automation 2017, an international conference held in Warsaw from March 15 to 17, 2017. It discusses research findings associated with the concepts behind INDUSTRY 4.0, with a focus on offering a better understanding of and promoting participation in the Fourth Industrial Revolution. Each chapter presents a detailed analysis of a specific technical problem, in most cases followed by a numerical analysis, simulation and description of the results of implementing the solution in a real-world context. The theoretical results, practical solutions and guidelines presented are valuable for both researchers working in the area of engineering sciences and practitioners looking for solutions to industrial problems. .

  4. Marketing automation

    Directory of Open Access Journals (Sweden)

    TODOR Raluca Dania

    2017-01-01

    Full Text Available The automation of the marketing process seems to be nowadays, the only solution to face the major changes brought by the fast evolution of technology and the continuous increase in supply and demand. In order to achieve the desired marketing results, businessis have to employ digital marketing and communication services. These services are efficient and measurable thanks to the marketing technology used to track, score and implement each campaign. Due to the technical progress, the marketing fragmentation, demand for customized products and services on one side and the need to achieve constructive dialogue with the customers, immediate and flexible response and the necessity to measure the investments and the results on the other side, the classical marketing approached had changed continue to improve substantially.

  5. Network-based automation for SMEs

    DEFF Research Database (Denmark)

    Parizi, Mohammad Shahabeddini; Radziwon, Agnieszka

    2017-01-01

    The implementation of appropriate automation concepts which increase productivity in Small and Medium Sized Enterprises (SMEs) requires a lot of effort, due to their limited resources. Therefore, it is strongly recommended for small firms to open up for the external sources of knowledge, which...... could be obtained through network interaction. Based on two extreme cases of SMEs representing low-tech industry and an in-depth analysis of their manufacturing facilities this paper presents how collaboration between firms embedded in a regional ecosystem could result in implementation of new...... with other members of the same regional ecosystem. The findings highlight two main automation related areas where manufacturing SMEs could leverage on external sources on knowledge – these are assistance in defining automation problem as well as appropriate solution and provider selection. Consequently...

  6. Automated Cryocooler Monitor and Control System Software

    Science.gov (United States)

    Britchcliffe, Michael J.; Conroy, Bruce L.; Anderson, Paul E.; Wilson, Ahmad

    2011-01-01

    This software is used in an automated cryogenic control system developed to monitor and control the operation of small-scale cryocoolers. The system was designed to automate the cryogenically cooled low-noise amplifier system described in "Automated Cryocooler Monitor and Control System" (NPO-47246), NASA Tech Briefs, Vol. 35, No. 5 (May 2011), page 7a. The software contains algorithms necessary to convert non-linear output voltages from the cryogenic diode-type thermometers and vacuum pressure and helium pressure sensors, to temperature and pressure units. The control function algorithms use the monitor data to control the cooler power, vacuum solenoid, vacuum pump, and electrical warm-up heaters. The control algorithms are based on a rule-based system that activates the required device based on the operating mode. The external interface is Web-based. It acts as a Web server, providing pages for monitor, control, and configuration. No client software from the external user is required.

  7. Electroconvulsive therapy in patients with cardiac pacemakers and implantable cardioverter defibrillators.

    Science.gov (United States)

    Dolenc, Tamara J; Barnes, Roxann D; Hayes, David L; Rasmussen, Keith G

    2004-09-01

    Electroconvulsive therapy (ECT) is used to treat major depressive illness, especially in elderly and medically frail patients. Not uncommonly, these patients have cardiac pacemakers or implantable cardioverter defibrillators (ICDs). Only a few case reports in the literature describe the use of ECT in such patients. Herein we review our ECT experience treating 26 pacemaker patients and 3 ICD patients. All patients obtained significant antidepressant benefits with ETC. Only one serious cardiac event occurred, a case of supraventricular tachycardia (SVT) requiring a stay on the cardiac intensive care unit. The SVT resolved and the patient went on to receive further uncomplicated ECT treatments. We conclude from this experience that with proper pre-ECT cardiac and pacemaker/defibrillator assessment, ECT can be safely and effectively administered to patients with an implanted cardiac device.

  8. Intensity of primary emotions in patients after implantation of an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Stoier, Louise; Pedersen, Preben Ulrich; Berg, Selina Kikkenborg

    2013-01-01

    Background: Experienced emotions can affect the outcome of, and adherence to a cardiac rehabilitation program, and patients coping with an illness. With more awareness of the expressed emotions, health professionals might be better able to understand the reactions of patients and to improve...... the support needed for coping. Living with an Implantable Cardi- overter Defibrillator can lead to anxiety and depression. Focus on the intensity of the primary emotions might be a potential to prevent development of these psychological states. Objectives: The aim of this paper are 1) to describe...... the intensity of primary emotions in patients after implantation of an Implantable Cardioverter Defibrillator and 2) to compare them with both the intensity of primary emotions in patients with a recent Myocardial Infarction and with a healthy population. Method: The intensity of primary emotions in patients...

  9. Electromagnetic Interference in Patients with Implanted Cardioverter-Defibrillators and Implantable Loop Recorders

    Directory of Open Access Journals (Sweden)

    Marcos de Sousa

    2002-07-01

    Full Text Available Modern life exposes us all to an ever-increasing number of potential sources of electromagnetic interference (EMI and patients with Implantable rhythm devices (IRD like pacemakers, implantable cardioverter defibrillators or implantable loop recorders often ask about the use of microwave ovens, walking through airport metal detectors and the use of cellular phones. Electromagnetic interference occurs when electromagnetic waves emitted by one device impede the normal function of another electronic device. The potential for interaction between implanted pacing systems and cardioverter-defibrillators (electromagnetic interference, EMI has been recognized for years.1,2,3,4. It has been shown that EMI can produce clinically significant effects on patients with implanted pacemakers and ICDs. For these reasons the following text discusses the influence of several EMI generating devices on IRD .

  10. Wearable cardioverter defibrillator: A life vest till the life boat (ICD arrives

    Directory of Open Access Journals (Sweden)

    Johnson Francis

    2014-01-01

    Full Text Available It is well established that implantable cardioverter defibrillator (ICD is a life saving device ensuring protection against life threatening ventricular arrhythmias. But there are certain situations like a recent myocardial infarction where the standard guidelines do not recommend the implantation of an ICD while the patient can still be at a risk of demise due to a life threatening ventricular arrhythmia. There could also be a temporary indication for protection while explanting an infected ICD system. The wearable cardioverter defibrillator (WCD is a device which comes to the rescue in such situations. In this brief review, we discuss the historical aspects of the development of a WCD, technical aspects as well as the clinical trial data and real world scenario of its use.

  11. Infectious endocardial intracardiac defibrillator lead, infectious pericarditis, and delayed constrictive pericarditis

    Directory of Open Access Journals (Sweden)

    Mohsen Mir Mohammad Sadeghi

    2013-01-01

    Full Text Available The usage of Implantable Cardiac Defibrillator (ICD since 1980s is becoming more popular these days. The rate of both, endocarditis and constrictive pericarditis are low but it still needs attention. We are reporting a rare case of ICD endocarditis as a result of toe infection in a diabetic patient. This was followed by infectious pericarditis after device removal by open heart surgery and then delayed constrictive pericarditis.

  12. Beam Profile Disturbances from Implantable Pacemakers or Implantable Cardioverter-Defibrillator Interactions

    Energy Technology Data Exchange (ETDEWEB)

    Gossman, Michael S., E-mail: mgossman@tsrcc.com [Tri-State Regional Cancer Center, Medical Physics Section, Ashland, KY (United States); Comprehensive Heart and Vascular Associates, Heart and Vascular Center, Ashland, KY (United States); Medtronic, Inc., External Research Program, Mounds View, MN (United States); Nagra, Bipinpreet; Graves-Calhoun, Alison; Wilkinson, Jeffrey [Tri-State Regional Cancer Center, Medical Physics Section, Ashland, KY (United States); Comprehensive Heart and Vascular Associates, Heart and Vascular Center, Ashland, KY (United States); Medtronic, Inc., External Research Program, Mounds View, MN (United States)

    2011-01-01

    The medical community is advocating for progressive improvement in the design of implantable cardioverter-defibrillators and implantable pacemakers to accommodate elevations in dose limitation criteria. With advancement already made for magnetic resonance imaging compatibility in some, a greater need is present to inform the radiation oncologist and medical physicist regarding treatment planning beam profile changes when such devices are in the field of a therapeutic radiation beam. Treatment plan modeling was conducted to simulate effects induced by Medtronic, Inc.-manufactured devices on therapeutic radiation beams. As a continuation of grant-supported research, we show that radial and transverse open beam profiles of a medical accelerator were altered when compared with profiles resulting when implantable pacemakers and cardioverter-defibrillators are placed directly in the beam. Results are markedly different between the 2 devices in the axial plane and the sagittal planes. Vast differences are also presented for the therapeutic beams at 6-MV and 18-MV x-ray energies. Maximum changes in percentage depth dose are observed for the implantable cardioverter-defibrillator as 9.3% at 6 MV and 10.1% at 18 MV, with worst distance to agreement of isodose lines at 2.3 cm and 1.3 cm, respectively. For the implantable pacemaker, the maximum changes in percentage depth dose were observed as 10.7% at 6 MV and 6.9% at 18 MV, with worst distance to agreement of isodose lines at 2.5 cm and 1.9 cm, respectively. No differences were discernible for the defibrillation leads and the pacing lead.

  13. Spiritual well-being may buffer psychological distress in patients with implantable cardioverter defibrillators (ICD)

    OpenAIRE

    Salmoirago-Blotcher, Elena; Crawford, Sybil; Tran, Chau; Goldberg, Robert; Rosenthal, Lawrence; Ockene, Ira

    2012-01-01

    Psychological distress is common in patients with implantable cardioverter defibrillators (ICDs) and has been associated with a worse prognosis. The authors examined whether spiritual wellbeing is associated with reduced psychological distress in patients with ICDs. The Functional Assessment of Chronic Illness Therapy-Spiritual Wellbeing (FACIT-SWB) questionnare and the Hospital Anxiety and Depression Scale (HADS) were used to measure spiritual wellbeing and overall psychological distress. Mu...

  14. Both Automation and Paper.

    Science.gov (United States)

    Purcell, Royal

    1988-01-01

    Discusses the concept of a paperless society and the current situation in library automation. Various applications of automation and telecommunications are addressed, and future library automation is considered. Automation at the Monroe County Public Library in Bloomington, Indiana, is described as an example. (MES)

  15. Frequent Home Monitoring of ICD Is Effective to Prevent Inappropriate Defibrillator Shock Delivery

    Directory of Open Access Journals (Sweden)

    Paolo Bifulco

    2014-01-01

    Full Text Available Recently, in the context of telemedicine, telemonitoring services are gaining attention. They are offered, for example, to patients with implantable cardioverter defibrillators (ICDs. A major problem associated with ICD therapy is the occurrence of inappropriate shocks which impair patients’ quality of life and may also be arrhythmogenic. The telemonitoring can provide a valid support to intensify followup visits, in order to improve the prevention of inappropriate defibrillator shock, thus enhancing patient safety. Inappropriate shock generally depends on atrial fibrillation, supraventricular tachycardia, and abnormal sensing (such as those caused by electromagnetic interferences. As a practical example, an unusual case of an ICD patient who risked an inappropriate shock while taking a shower is reported. Continuous remote telemonitoring was able to timely warn cardiologist via GSM-SMS, who were able to detect improper sensing examining the intracardiac electrogram via Web. Patient was promptly contacted and warned to not further come in contact with the hydraulic system and any electrical appliance to prevent an inappropriate defibrillator shock. This demonstrates the effectiveness and usefulness of continuous remote telemonitoring in supporting ICD patients.

  16. The use of guideline recommended beta-blocker therapy in primary prevention implantable cardioverter defibrillator patients

    DEFF Research Database (Denmark)

    Ruwald, Anne Christine; Gislason, Gunnar Hilmar; Vinther, Michael

    2017-01-01

    Aims: We aimed to examine the use of guideline recommended beta-blocker therapy prior to and after primary prevention implantable cardioverter defibrillator (ICD) implantation in a 'real-life' setting. Methods and results: From the Danish Pacemaker and ICD Registry we identified all 1st-time prim......Aims: We aimed to examine the use of guideline recommended beta-blocker therapy prior to and after primary prevention implantable cardioverter defibrillator (ICD) implantation in a 'real-life' setting. Methods and results: From the Danish Pacemaker and ICD Registry we identified all 1st......-time primary prevention ICD and cardiac resynchronization therapy defibrillator (CRT-D) implantations in Denmark from 2007-12 (n = 2935). Use of beta-blocker, type and dose was acquired through the Danish Prescription Registry. According to guideline recommendations, we defined target daily doses as ≥50 mg...... carvedilol and ≥200 mg metoprolol. Prior to implantation 2427 of 2935 (83%) patients received beta-blocker therapy, with 2166 patients (89%) having initiated treatment 3 months or more prior to implantation. The majority of patients was prescribed carvedilol (52%) or metoprolol (41%). Patients on carvedilol...

  17. Cardiopulmonary Resuscitation (CPR): First Aid

    Science.gov (United States)

    ... learn CPR properly, take an accredited first-aid training course, including CPR and how to use an automated external defibrillator ( ... and Research. © 1998-2018 Mayo Foundation for Medical Education and Research (MFMER). All rights reserved.

  18. Ventricular Fibrillation

    Science.gov (United States)

    ... to the heart with a device called an automated external defibrillator (AED). Treatments to prevent sudden cardiac ... number of places, such as in airplanes, police cars and shopping malls. They can even be purchased ...

  19. CPR - infant

    Science.gov (United States)

    ... as 4 to 6 minutes later. Machines called automated external defibrillators (AEDs) can be found in many ... side down. Follow the guidelines for using infant car seats. Teach your baby the meaning of "don' ...

  20. Radiotherapy and risk of implantable cardioverter-defibrillator malfunctions: experimental data from direct exposure at increasing doses.

    Science.gov (United States)

    Zecchin, Massimo; Artico, Jessica; Morea, Gaetano; Severgnini, Mara; Bianco, Elisabetta; De Luca, Antonio; Fantasia, Anna Zorzin; Salvatore, Luca; Milan, Vittorino; Lucarelli, Matteo; Dissegna, Roberta; Cannatà, Antonio; Sinagra, Gianfranco

    2018-04-01

    During radiotherapy, in patients with implantable cardioverter-defibrillators (ICDs) malfunctions are considered more likely if doses more than 2 Gy reach the ICD site; however, most malfunctions occur with high-energy (>10 MV) radiations, and the risk is less defined using 6-MV linear accelerators. The purpose of the study is to experimentally evaluate the occurrence of malfunctions in ICDs radiated with a 6-MV linear accelerator at increasing photon doses. Thirty-two ICDs from all manufacturers (31 explanted and one demo) were evaluated; all devices with a sufficient battery charge underwent multiple radiations with a 6-MV photon beam reaching a cumulative dose at ICD site of 0.5, 1, 2, 3, 5 and 10 Gy and interrogated after every session. All antitachycardia therapies were left enabled; two ICDs were connected to a rhythm simulator (one simulating a complete atrioventricular block without ventricular activity) and visually monitored by external ECG and the ICD programmer during radiation. Thirteen ICDs were excluded before radiation because of battery depletion; after radiation up to the cumulative dose at the cardiac implantable electronic device site of 10 Gy, in the remaining 19 devices, programmation and battery charge remained unchanged and no switch to safety mode was observed; oversensing, pacing inhibition or inappropriate antitachycardia therapy were neither recorded nor visually observed during radiation. With a low-energy accelerator, neither malfunctions nor electromagnetic interferences were detected radiating the ICDs at doses usually reaching the ICD pocket during radiotherapy sessions. In this context, magnet application to avoid oversensing and inappropriate therapy seems, therefore, useless.

  1. Regulating multiple externalities

    DEFF Research Database (Denmark)

    Waldo, Staffan; Jensen, Frank; Nielsen, Max

    2016-01-01

    Open access is a well-known externality problem in fisheries causing excess capacity and overfishing. Due to global warming, externality problems from CO2 emissions have gained increased interest. With two externality problems, a first-best optimum can be achieved by using two regulatory instrume......Open access is a well-known externality problem in fisheries causing excess capacity and overfishing. Due to global warming, externality problems from CO2 emissions have gained increased interest. With two externality problems, a first-best optimum can be achieved by using two regulatory...

  2. Comparison of low-energy versus high-energy biphasic defibrillation shocks following prolonged ventricular fibrillation.

    Science.gov (United States)

    Walcott, Gregory P; Melnick, Sharon B; Killingsworth, Cheryl R; Ideker, Raymond E

    2010-01-01

    Since the initial development of the defibrillator, there has been concern that, while delivery of a large electric shock would stop fibrillation, it would also cause damage to the heart. This concern has been raised again with the development of the biphasic defibrillator. To compare defibrillation efficacy, postshock cardiac function, and troponin I levels following 150-J and 360-J shocks. Nineteen swine were anesthetized with isoflurane and instrumented with pressure catheters in the left ventricle, aorta, and right atrium. The animals were fibrillated for 6 minutes, followed by defibrillation with either low-energy (n = 8) or high-energy (n = 11) shocks. After defibrillation, chest compressions were initiated and continued until return of spontaneous circulation (ROSC). Epinephrine, 0.01 mg/kg every 3 minutes, was given for arterial blood pressure < 50 mmHg. Hemodynamic parameters were recorded for four hours. Transthoracic echocardiography was performed and troponin I levels were measured at baseline and four hours following ventricular fibrillation (VF). Survival rates at four hours were not different between the two groups (low-energy, 5 of 8; high-energy, 7 of 11). Results for arterial blood pressure, positive dP/dt (first derivative of pressure measured over time, a measure of left ventricular contractility), and negative dP/dt at the time of lowest arterial blood pressure (ABP) following ROSC were not different between the two groups (p = not significant [NS]), but were lower than at baseline. All hemodynamic measures returned to baseline by four hours. Ejection fractions, stroke volumes, and cardiac outputs were not different between the two groups at four hours. Troponin I levels at four hours were not different between the two groups (12 +/- 11 ng/mL versus 21 +/- 26 ng/mL, p = NS) but were higher at four hours than at baseline (19 +/- 19 ng/mL versus 0.8 +/- 0.5 ng/mL, p < 0.05, groups combined). Biphasic 360-J shocks do not cause more cardiac damage

  3. Non-linear dynamical signal characterization for prediction of defibrillation success through machine learning

    Directory of Open Access Journals (Sweden)

    Shandilya Sharad

    2012-10-01

    Full Text Available Abstract Background Ventricular Fibrillation (VF is a common presenting dysrhythmia in the setting of cardiac arrest whose main treatment is defibrillation through direct current countershock to achieve return of spontaneous circulation. However, often defibrillation is unsuccessful and may even lead to the transition of VF to more nefarious rhythms such as asystole or pulseless electrical activity. Multiple methods have been proposed for predicting defibrillation success based on examination of the VF waveform. To date, however, no analytical technique has been widely accepted. We developed a unique approach of computational VF waveform analysis, with and without addition of the signal of end-tidal carbon dioxide (PetCO2, using advanced machine learning algorithms. We compare these results with those obtained using the Amplitude Spectral Area (AMSA technique. Methods A total of 90 pre-countershock ECG signals were analyzed form an accessible preshosptial cardiac arrest database. A unified predictive model, based on signal processing and machine learning, was developed with time-series and dual-tree complex wavelet transform features. Upon selection of correlated variables, a parametrically optimized support vector machine (SVM model was trained for predicting outcomes on the test sets. Training and testing was performed with nested 10-fold cross validation and 6–10 features for each test fold. Results The integrative model performs real-time, short-term (7.8 second analysis of the Electrocardiogram (ECG. For a total of 90 signals, 34 successful and 56 unsuccessful defibrillations were classified with an average Accuracy and Receiver Operator Characteristic (ROC Area Under the Curve (AUC of 82.2% and 85%, respectively. Incorporation of the end-tidal carbon dioxide signal boosted Accuracy and ROC AUC to 83.3% and 93.8%, respectively, for a smaller dataset containing 48 signals. VF analysis using AMSA resulted in accuracy and ROC AUC of 64

  4. ExternE National Implementation Finland

    Energy Technology Data Exchange (ETDEWEB)

    Pingoud, K; Maelkki, H; Wihersaari, M; Pirilae, P [VTT Energy, Espoo (Finland); Hongisto, M [Imatran Voima Oy, Vantaa (Finland); Siitonen, S [Ekono Energy Ltd, Espoo (Finland); Johansson, M [Finnish Environment Institute, Helsinki (Finland)

    1999-07-01

    ExternE National Implementation is a continuation of the ExternE Project, funded in part by the European Commission's Joule III Programme. This study is the result of the ExternE National Implementation Project for Finland. Three fuel cycles were selected for the Finnish study: coal, peat and wood-derived biomass, which together are responsible for about 40% of total electricity generation in Finland and about 75% of the non-nuclear fuel based generation. The estimated external costs or damages were dominated by the global warming (GW) impacts in the coal and peat fuel cycles, but knowledge of the true GW impacts is still uncertain. From among other impacts that were valued in monetary terms the human health damages due to airborne emissions dominated in all the three fuel cycles. Monetary valuation for ecosystem impacts is not possible using the ExternE methodology at present. The Meri-Pori power station representing the coal fuel cycle is one of the world's cleanest and most efficient coal-fired power plants with a condensing turbine. The coal is imported mainly from Poland. The estimated health damages were about 4 mECU/kWh, crop damages an order of magnitude lower and damages caused to building materials two orders of magnitude lower. The power stations of the peat and biomass fuel cycles are of CHP type, generating electricity and heat for the district heating systems of two cities. Their fuels are of domestic origin. The estimated health damages allocated to electricity generation were about 5 and 6 mECU/kWh, respectively. The estimates were case-specific and thus an generalisation of the results to the whole electricity generation in Finland is unrealistic. Despite the uncertainties and limitations of the methodology, it is a promising tool in the comparison of similar kinds of fuel cycles, new power plants and pollution abatement technologies and different plant locations with each other. (orig.)

  5. ExternE National Implementation Finland

    International Nuclear Information System (INIS)

    Pingoud, K.; Maelkki, H.; Wihersaari, M.; Pirilae, P.; Hongisto, M.; Siitonen, S.; Johansson, M.

    1999-01-01

    ExternE National Implementation is a continuation of the ExternE Project, funded in part by the European Commission's Joule III Programme. This study is the result of the ExternE National Implementation Project for Finland. Three fuel cycles were selected for the Finnish study: coal, peat and wood-derived biomass, which together are responsible for about 40% of total electricity generation in Finland and about 75% of the non-nuclear fuel based generation. The estimated external costs or damages were dominated by the global warming (GW) impacts in the coal and peat fuel cycles, but knowledge of the true GW impacts is still uncertain. From among other impacts that were valued in monetary terms the human health damages due to airborne emissions dominated in all the three fuel cycles. Monetary valuation for ecosystem impacts is not possible using the ExternE methodology at present. The Meri-Pori power station representing the coal fuel cycle is one of the world's cleanest and most efficient coal-fired power plants with a condensing turbine. The coal is imported mainly from Poland. The estimated health damages were about 4 mECU/kWh, crop damages an order of magnitude lower and damages caused to building materials two orders of magnitude lower. The power stations of the peat and biomass fuel cycles are of CHP type, generating electricity and heat for the district heating systems of two cities. Their fuels are of domestic origin. The estimated health damages allocated to electricity generation were about 5 and 6 mECU/kWh, respectively. The estimates were case-specific and thus an generalisation of the results to the whole electricity generation in Finland is unrealistic. Despite the uncertainties and limitations of the methodology, it is a promising tool in the comparison of similar kinds of fuel cycles, new power plants and pollution abatement technologies and different plant locations with each other. (orig.)

  6. ExternE National Implementation Finland

    Energy Technology Data Exchange (ETDEWEB)

    Pingoud, K.; Maelkki, H.; Wihersaari, M.; Pirilae, P. [VTT Energy, Espoo (Finland); Hongisto, M. [Imatran Voima Oy, Vantaa (Finland); Siitonen, S. [Ekono Energy Ltd, Espoo (Finland); Johansson, M. [Finnish Environment Institute, Helsinki (Finland)

    1999-07-01

    ExternE National Implementation is a continuation of the ExternE Project, funded in part by the European Commission's Joule III Programme. This study is the result of the ExternE National Implementation Project for Finland. Three fuel cycles were selected for the Finnish study: coal, peat and wood-derived biomass, which together are responsible for about 40% of total electricity generation in Finland and about 75% of the non-nuclear fuel based generation. The estimated external costs or damages were dominated by the global warming (GW) impacts in the coal and peat fuel cycles, but knowledge of the true GW impacts is still uncertain. From among other impacts that were valued in monetary terms the human health damages due to airborne emissions dominated in all the three fuel cycles. Monetary valuation for ecosystem impacts is not possible using the ExternE methodology at present. The Meri-Pori power station representing the coal fuel cycle is one of the world's cleanest and most efficient coal-fired power plants with a condensing turbine. The coal is imported mainly from Poland. The estimated health damages were about 4 mECU/kWh, crop damages an order of magnitude lower and damages caused to building materials two orders of magnitude lower. The power stations of the peat and biomass fuel cycles are of CHP type, generating electricity and heat for the district heating systems of two cities. Their fuels are of domestic origin. The estimated health damages allocated to electricity generation were about 5 and 6 mECU/kWh, respectively. The estimates were case-specific and thus an generalisation of the results to the whole electricity generation in Finland is unrealistic. Despite the uncertainties and limitations of the methodology, it is a promising tool in the comparison of similar kinds of fuel cycles, new power plants and pollution abatement technologies and different plant locations with each other. (orig.)

  7. External Otitis (Swimmer's Ear)

    Science.gov (United States)

    ... otitis. Fungal external otitis (otomycosis), typically caused by Aspergillus niger or Candida albicans, is less common. Boils are ... in the ear. Fungal external otitis caused by Aspergillus niger usually causes grayish black or yellow dots (called ...

  8. ExternE: Externalities of energy Vol. 2. Methodology

    International Nuclear Information System (INIS)

    Berry, J.; Holland, M.; Watkiss, P.

    1995-01-01

    This report describes the methodology used by the ExternE Project of the European Commission (DGXII) JOULE Programme for assessment of the external costs of energy. It is one of a series of reports describing analysis of nuclear, fossil and renewable fuel cycles for assessment of the externalities associated with electricity generation. Part I of the report deals with analysis of impacts, and Part II with the economic valuation of those impacts. Analysis is conducted on a marginal basis, to allow the effect of an incremental investment in a given technology to be quantified. Attention has been paid to the specificity of results with respect to the location of fuel cycle activities, the precise technologies used, and the type and source of fuel. The main advantages of this detailed approach are as follows: It takes full and proper account of the variability of impacts that might result from different power projects; It is more transparent than analysis based on hypothetically 'representative' cases for each of the different fuel cycles; It provides a framework for consistent comparison between fuel cycles. A wide variety of impacts have been considered. These include the effects of air pollution on the natural and human environment, consequences of accidents in the workplace, impacts of noise and visual intrusion on amenity, and the effects of climate change arising from the release of greenhouse gases. Wherever possible we have used the 'impact pathway' or 'damage function' approach to follow the analysis from identification of burdens (e.g. emissions) through to impact assessment and then valuation in monetary terms. This has required a detailed knowledge of the technologies involved, pollutant dispersion, analysis of effects on human and environmental health, and economics. In view of this the project brought together a multi-disciplinary team with experts from many European countries and the USA. The spatial and temporal ranges considered in the analysis are

  9. del Nido versus St. Thomas Cardioplegia Solutions: A Single-Center Retrospective Analysis of Post Cross-Clamp Defibrillation Rates.

    Science.gov (United States)

    Buel, Shane T; Striker, Carrie Whittaker; O'Brien, James E

    2016-06-01

    There are many cardioplegia solutions currently in use for pediatric cardiopulmonary bypass (CPB). The most common being del Nido solution. Another common cardioplegia solution used for pediatric CPB is St. Thomas. In October 2014, Children's Mercy Kansas City changed from the use of modified St. Thomas to del Nido. This study compared rates of post cross-clamp fibrillation requiring defibrillation between del Nido solution and modified St. Thomas solution stratified by weight at Children's Mercy Kansas City. This retrospective study consisted of 394 patients who underwent cardiac surgery requiring cardioplegia between January 1, 2014 and July 31, 2015. The outcome measured was defibrillation upon cross-clamp removal. Statistical significance was determined using Fishers exact test with a two-sided significance level of .05. Incidence of defibrillation post cross-clamp removal was 4.4% in the del Nido group and 26.8% in the St. Thomas group (p Thomas group (p Thomas group (p Thomas group (p 60-kg category had an incidence of defibrillation of 16.7% in the del Nido group and 63% in the St. Thomas group (p Thomas and del Nido cardioplegia solutions. Analyses of weight stratifications demonstrate a decrease in the rate of defibrillation post cross-clamp removal in all categories within the del Nido group.

  10. ASH External Web Portal (External Portal) -

    Data.gov (United States)

    Department of Transportation — The ASH External Web Portal is a web-based portal that provides single sign-on functionality, making the web portal a single location from which to be authenticated...

  11. Symptomatic heart failure is the most important clinical correlate of impaired quality of life, anxiety, and depression in implantable cardioverter-defibrillator patients

    DEFF Research Database (Denmark)

    Johansen, Jens B; Pedersen, Susanne S.; Spindler, Helle

    2008-01-01

    To identify correlates of impaired quality of life (QOL), anxiety, and depression in patients with an implantable cardioverter-defibrillator (ICD).......To identify correlates of impaired quality of life (QOL), anxiety, and depression in patients with an implantable cardioverter-defibrillator (ICD)....

  12. 41 CFR 102-79.115 - What guidelines must an agency follow if it elects to establish a public access defibrillation...

    Science.gov (United States)

    2010-07-01

    ... SPACE Assignment and Utilization of Space Public Access Defibrillation Programs § 102-79.115 What... 41 Public Contracts and Property Management 3 2010-07-01 2010-07-01 false What guidelines must an agency follow if it elects to establish a public access defibrillation program in a Federal facility? 102...

  13. Pacemaker System Malfunction Resulting from External Electrical Cardioversion: A Case Report

    Directory of Open Access Journals (Sweden)

    Taku Nishida, MD

    2009-01-01

    Full Text Available In May 2005 a 68-year-old woman received a VDD pacemaker implantation in the right pectoral region at our hospital for the treatment of complete atrioventricular block. In July 2008, she was diagnosed with dilated cardiomyopathy based on histological testing. In November 2008, she developed syncope due to ventricular tachycardia while at another hospital. She underwent external electrical cardioversion with an anterior-lateral paddle position using a single shock of 100 J. This shock led to severe bradycardia resulting in a transfer to our hospital. The physician who provided the shock could not have been aware that the patient had an implanted pacemaker. The skin above the pulse generator was burned. The electrocardiogram showed no pacing spikes or ventricular escape rhythm. Investigation of the pacemaker 3 hours after cardioversion revealed reprogramming of the device and a marked rise in the lead impedance (>3,000 ohm. Removal of the generator and implantation of a biventricular cardioverter defibrillator were required. The emergency situation, the small size of the generator, the small incision made using the buried suture method, and the patient's obesity all probably contributed to the physician's not noticing the implanted pacemaker. It is important to increase awareness of the severe consequences that may follow if the physician administering external defibrillation does not know about the patient's implanted pacemaker.

  14. An automated swimming respirometer

    DEFF Research Database (Denmark)

    STEFFENSEN, JF; JOHANSEN, K; BUSHNELL, PG

    1984-01-01

    An automated respirometer is described that can be used for computerized respirometry of trout and sharks.......An automated respirometer is described that can be used for computerized respirometry of trout and sharks....

  15. Autonomy and Automation

    Science.gov (United States)

    Shively, Jay

    2017-01-01

    A significant level of debate and confusion has surrounded the meaning of the terms autonomy and automation. Automation is a multi-dimensional concept, and we propose that Remotely Piloted Aircraft Systems (RPAS) automation should be described with reference to the specific system and task that has been automated, the context in which the automation functions, and other relevant dimensions. In this paper, we present definitions of automation, pilot in the loop, pilot on the loop and pilot out of the loop. We further propose that in future, the International Civil Aviation Organization (ICAO) RPAS Panel avoids the use of the terms autonomy and autonomous when referring to automated systems on board RPA. Work Group 7 proposes to develop, in consultation with other workgroups, a taxonomy of Levels of Automation for RPAS.

  16. Configuration Management Automation (CMA) -

    Data.gov (United States)

    Department of Transportation — Configuration Management Automation (CMA) will provide an automated, integrated enterprise solution to support CM of FAA NAS and Non-NAS assets and investments. CMA...

  17. The oral cavity is not a primary source for implantable pacemaker or cardioverter defibrillator infections

    Science.gov (United States)

    2013-01-01

    Background To test the hypothesis that the oral cavity is a potential source for implantable pacemaker and cardioverter defibrillators infections, the bacterial diversity on explanted rhythm heart management devices was investigated and compared to the oral microbiome. Methods A metagenomic approach was used to analyze the bacterial diversity on the surfaces of non-infected and infected pacemakers. The DNA from surfaces swaps of 24 non-infected and 23 infected pacemaker were isolated and subjected to bacterial-specific DNA amplification, single strand conformation polymorphism- (SSCP) and sequencing analysis. Species-specific primer sets were used to analyze for any correlation between bacterial diversity on pacemakers and in the oral cavity. Results DNA of bacterial origin was detected in 21 cases on infected pacemakers and assigned to the bacterial phylotypes Staphylococcus epidermidis, Propionibacterium acnes, Staphylococcus aureus, Staphylococcus schleiferi and Stapyhlococcus. In 17 cases bacterial DNA was found on pacemakers with no clinical signs of infections. On the basis of the obtained sequence data, the phylotypes Propionibacterium acnes, Staphylococcus and an uncultured bacterium were identified. Propionibacterium acnes and Staphylococcus epidermidis were the only bacteria detected in pacemeaker (n = 25) and oral samples (n = 11). Conclusions The frequency of the coincidental detection of bacteria on infected devices and in the oral cavity is low and the detected bacteria are highly abundant colonizers of non-oral human niches. The transmission of oral bacteria to the lead or device of implantable pacemaker or cardioverter defibrillators is unlikely relevant for the pathogenesis of pacemaker or cardioverter defibrillators infections. PMID:23575037

  18. ExternE: Externalities of energy Vol. 1. Summary

    International Nuclear Information System (INIS)

    Holland, M.; Berry, J.

    1995-01-01

    There is a growing requirement for policy analysts to take account of the environment in their decision making and to undertake the specified cost-benefit analysis. Within the European Union this is reflected in the 5th Environmental Action Programme, and the Commission's White Paper entitled 'Growth, competitiveness, employment and the ways forward to the 21st century'. This has led to a need for evaluation of environmental externalities. The ExternE Project commenced in 1991 as the European part of a collaborative study between the European Commission and the US Department of Energy. It aims to be the first systematic approach to the evaluation of external costs of a wide range of different fuel cycles. The project will result in an operational accounting framework for the quantification and monetarisation of priority environmental and other externalities. This framework will allow the calculation of the marginal external costs and benefits for specific power plants, at specific sites using specified technologies. There are three major phases in the project. Phase 1 was undertaken in collaboration with the US Department of Energy. In this phase the teams jointly developed the conceptual approach and methodology and shared scientific information for application to a number of fuel cycles. On the European side work concentrated on the nuclear and coal fuel cycles which together were expected to raise many of the fundamental issues in fuel cycle analysis. The project is currently nearing completion of Phase 2. During this phase the methodology has been applied to a wide range of different fossil, nuclear and renewable fuel cycles for power generation and energy conservation options. Also a series of National Implementation Programmes is underway in which the methodology and accounting framework are being applied to reference sites throughout Europe. In addition the general methodology is being extended to address the evaluation of externalities associated with

  19. Externalities of fuel cycles 'ExternE' project. Summary report

    International Nuclear Information System (INIS)

    Holland, M.; Berry, J.

    1994-01-01

    There is a growing requirement for policy analysts to take account of the environment in their decision making and to undertake the specified cost-benefit analysis. Within the European Union this is reflected in the 5th Environmental Action Programme, and the Commission's White Paper entitled 'Growth, competitiveness, employment and the ways forward to the 21st century'. This has led to a need for evaluation of environmental externalities. The ExternE Project commenced in 1991 as the European part of a collaborative study between the European Commission and the US Department of Energy. It aims to be the first systematic approach to the evaluation of external costs of a wide range of different fuel cycles. The project will result in an operational accounting framework for the quantification and monetarisation of priority environmental and other externalities. This framework will allow the calculation of the marginal external costs and benefits for specific power plants, at specific sites using specified technologies. There are three major phases in the project. Phase I was undertaken in collaboration with the US Department of Energy. In this phase the teams jointly developed the conceptual approach and methodology and shared scientific information for application to a number of fuel cycles. On the European side work concentrated on the nuclear and coal fuel cycles which together were expected to raise many of the fundamental issues in fuel cycle analysis. The project is currently nearing completion of Phase 2. During this phase the methodology has been applied to a wide range of different fossil, nuclear and renewable fuel cycles for power generation and energy conservation options. Also a series of National Implementation Programmes are underway in which the methodology and accounting framework are being applied to reference sites throughout Europe. In addition the general methodology is being extended to address the evaluation of externalities associated with

  20. Venous Obstruction Following Pacemaker or Implantable Cardioverter-Defibrillator Implantation, Mini Review

    Directory of Open Access Journals (Sweden)

    Mohammad Ali Akbarzadeh

    2016-08-01

    Full Text Available Venous obstruction is relatively frequent following permanent pacemaker or implantable cardioverter-defibrillator (ICD implantation. However, most of them are asymptomatic. Although the exact risk factor for this complication is not known, number of leads, heart failure and infection may prone the patient to this complication. The goal standard for detection of vein stenosis is venography; however, ultrasound sonography has an acceptable accuracy. Anticoagulant therapy may be considered for symptomatic patients. For device upgrading, non-functional leads removal, venoplasty and rarely surgical treatment may be indicated.

  1. Successful Treatment of Refractory Electrical Storm With Landiolol After More Than 100 Electrical Defibrillations.

    Science.gov (United States)

    Kanamori, Kenta; Aoyagi, Takashi; Mikamo, Takashi; Tsutsui, Kenta; Kunishima, Tomoyuki; Inaba, Hideko; Hayami, Noriyuki; Murakawa, Yuji

    2015-01-01

    Electrical storm (ES) was observed in an 82-year-old man with recent myocardial infarction. Conventional therapy, including amiodarone, could not suppress the ES. After more than 100 electrical defibrillations, we were finally able to control the ES with the administration of landiolol. It is known that landiolol can inhibit ES. However, we hesitate to use landiolol in patients with low cardiac output. We would like to emphasize that careful use of landiolol should be considered in patients with refractory ES after myocardial infarction, although cardiac output is severely reduced.

  2. External radiation surveillance

    International Nuclear Information System (INIS)

    Antonio, E.J.

    1995-01-01

    This section of the 1994 Hanford Site Environmental Report describes how external radiation was measured, how surveys were performed, and the results of these measurements and surveys. External radiation exposure rates were measured at locations on and off the Hanford Site using thermoluminescent dosimeters (TLD). External radiation and contamination surveys were also performed with portable radiation survey instruments at locations on and around the Hanford Site

  3. External radiation surveillance

    Energy Technology Data Exchange (ETDEWEB)

    Antonio, E.J.

    1995-06-01

    This section of the 1994 Hanford Site Environmental Report describes how external radiation was measured, how surveys were performed, and the results of these measurements and surveys. External radiation exposure rates were measured at locations on and off the Hanford Site using thermoluminescent dosimeters (TLD). External radiation and contamination surveys were also performed with portable radiation survey instruments at locations on and around the Hanford Site.

  4. Automation in Clinical Microbiology

    Science.gov (United States)

    Ledeboer, Nathan A.

    2013-01-01

    Historically, the trend toward automation in clinical pathology laboratories has largely bypassed the clinical microbiology laboratory. In this article, we review the historical impediments to automation in the microbiology laboratory and offer insight into the reasons why we believe that we are on the cusp of a dramatic change that will sweep a wave of automation into clinical microbiology laboratories. We review the currently available specimen-processing instruments as well as the total laboratory automation solutions. Lastly, we outline the types of studies that will need to be performed to fully assess the benefits of automation in microbiology laboratories. PMID:23515547

  5. Automated imaging system for single molecules

    Science.gov (United States)

    Schwartz, David Charles; Runnheim, Rodney; Forrest, Daniel

    2012-09-18

    There is provided a high throughput automated single molecule image collection and processing system that requires minimal initial user input. The unique features embodied in the present disclosure allow automated collection and initial processing of optical images of single molecules and their assemblies. Correct focus may be automatically maintained while images are collected. Uneven illumination in fluorescence microscopy is accounted for, and an overall robust imaging operation is provided yielding individual images prepared for further processing in external systems. Embodiments described herein are useful in studies of any macromolecules such as DNA, RNA, peptides and proteins. The automated image collection and processing system and method of same may be implemented and deployed over a computer network, and may be ergonomically optimized to facilitate user interaction.

  6. Dutch outcome in implantable cardioverter-defibrillator therapy (DO-IT) : Registry design and baseline characteristics of a prospective observational cohort study to predict appropriate indication for implantable cardioverter-defibrillator

    NARCIS (Netherlands)

    Van Barreveld, M. (M.); M.G.W. Dijkgraaf (Marcel); Hulleman, M. (M.); L. Boersma (Lucas); P.P.H.M. Delnoy (Peter Paul); M. Meine (Mathias); Tuinenburg, A.E. (A. E.); D.A.M.J. Theuns (Dominic); P. van der Voort (Pepijn); G-J.P. Kimman (Geert-Jan); E. Buskens (Erik); Tijssen, J.P.G. (J. P.G.); Bruinsma, N. (N.); Verstraelen, T.E. (T. E.); A.H. Zwinderman (Ailko); Van Dessel, P.H.F.M. (P. H.F.M.); A.A.M. Wilde (Arthur)

    2017-01-01

    textabstractBackground Implantable cardioverter-defibrillators (ICDs) are widely used for the prevention of sudden cardiac death. At present, both clinical benefit and cost-effectiveness of ICD therapy in primary prevention patients are topics of discussion, as only a minority of these patients will

  7. ExternE: Externalities of energy Vol. 5. Nuclear

    International Nuclear Information System (INIS)

    Dreicer, M.; Tort, V.; Manen, P.

    1995-01-01

    Since the early 1970s, there has been increased interest in the environmental impacts that are caused by the generation of electricity. The comparative risk assessment studies at that time used mainly deaths and injuries as impact indicators. By the end of the 1980s studies changed to the assessment of the costs imposed on society and the environment that were not included in the market price of the energy produced, the so-called external costs. The preliminary studies that were published set the conceptual basis, grounded in neo-classical economics, for the valuation of the health and environmental impacts that could be assessed. As a consequence of the many questions raised by the methodologies employed by these early studies, Directorate General XII (DG XII) of the Commission of the European Communities established a collaborative research programme with the United States Department of Energy to identify an appropriate methodology for this type of work. Following the completion of this collaboration, the DG XII programme has continued as the ExternE project. The main objective of the work carried out at CEPN was to develop an impact pathway methodology for the nuclear fuel cycle that would be consistent with the methodologies developed for other fuel cycles, without loosing the nuclear-specific techniques required for a proper evaluation. In this way, comparisons between the different fuel cycles would be possible. This report presents the methodology and demonstration of the results in the context of the French nuclear fuel cycle. The United States team at Oak Ridge National Laboratory has previously issued a draft report on the results of their assessment. The French fuel cycle was broken down into 8 separate stages. Reference sites and 1990s technology were chosen to represent the total nuclear fuel cycle, as it exists today. In addition, the transportation of material between the sites was considered. The facilities are assessed for routine operation, except

  8. Is there an Optimal Shape of the Defibrillation Shock: Constant Current vs. Pulsed Biphasic Waveforms

    Directory of Open Access Journals (Sweden)

    Ivan Dotsinsky

    2013-04-01

    Full Text Available Three waveforms for transthoracic defibrillation are assessed and compared: the Pulsed Biphasic Waveform (PBW, the Rectilinear Biphasic Waveform (RBW, and the "lossless" constant current (LLCC pulses. Two indices are introduced: 1 kf = W/W0 - the ratio between the delivered energy W and the energy W0 of a rectangular pulse with the same duration and electric charge; 2 ηC = W/WC0 - the level of utilizing the initially loaded capacitor energy WC0. The envisioned comparative study shows that ηC index is favorable for both PBW and LLCC, while kf of both RBW and LLCC demonstrates advantage over the PBW in the range of small inter-electrode thoracic impedances below 80 Ω. Some design considerations are also discussed. The attractive LLCC concept needs large and heavy inductive coil to support the constant current amplitude, besides it is capable to induce strong electromagnetic influences due to the complex current control. The RBW technology controls the delivery of current through a series of internal resistors which are, however, a source of high heat losses. The PBW implements controlled duty cycle of high-frequency chopped pulses to adapt the energy delivery in respect of the patient impedance measured at the beginning of the shock. PBW technology makes use of small capacitors which allows the construction of light weight and small-size portable devices for transthoracic defibrillation.

  9. How does an implantable cardioverter defibrillator (ICD) affect the lives of patients and their families?

    Science.gov (United States)

    Eckert, Marion; Jones, Tina

    2002-06-01

    This study aimed to identify the lived experience of patients with implantable cardioverter defibrillators (ICD) and their families. The methodology used was interpretative phenomenology. Unstructured interviews were conducted with three family members and three ICD recipients. Using a methodological approach outlined by van Manen, the participants transcribed texts were analysed looking for similar concepts and ideas that developed into themes that explicated the meaning of this phenomena. The themes that emerged were: dependence, which encompassed their perceptions about the life-saving device; the memory of their first defibrillation experience; lifestyle changes, which incorporated modification techniques; lack of control, which highlighted feelings such as fear, anxiety and powerlessness; mind game, which illustrated psychological challenges; and the issue of security, demonstrating how 'being there' and not 'being there' impacted on their everyday lives. The long-term outcomes of living with an ICD are important considerations for all health-care providers. This research highlights the everyday activities of recipients, the lifestyle changes they have made, the emotional significance of the device and the psychological coping strategies that the participants have adopted. The findings of this research will allow health-care professionals to be better prepared to provide education and support for ICD recipients and their families in regards to issues related to insertion of the device during the postinsertion recovery period and for long-term management after hospital discharge.

  10. Subclavian Vein Stenosis/Occlusion Following Transvenous Cardiac Pacemaker and Defibrillator Implantation: Incidence, Pathophysiology and Current Management

    Directory of Open Access Journals (Sweden)

    Brian O'Leary

    2015-08-01

    Full Text Available Subclavian vein stenosis is a common, but usually asymptomatic, complication following cardiac device placement. In addition to reviewing the literature on incidence, pathogenesis and management options for this important clinical problem, we describe two cases of symptomatic subclavian vein occlusion following pacemaker/defibrillator placement and successful treatment with venoplasty and stenting.

  11. A review of economic evaluation models for cardiac resynchronization therapy with implantable cardioverter defibrillators in patients with heart failure

    NARCIS (Netherlands)

    Tomini, F.; van Asselt, A. D.

    OBJECTIVES: Cardiac resynchronization therapy with biventricular pacemaker (CRT-P) is considered an effective treatment for heart failure (HF). Adding implantable cardioverter defibrillators (CRT-D) may further reduce the risk of sudden cardiac death (SCD). However, economic evaluations have shown

  12. A review of economic evaluation models for cardiac resynchronization therapy with implantable cardioverter defibrillators in patients with heart failure

    NARCIS (Netherlands)

    Tomini, F.; van Asselt, A. D.

    2012-01-01

    OBJECTIVES: Cardiac resynchronization therapy with biventricular pacemaker (CRT-P) is considered an effective treatment for heart failure (HF). Adding implantable cardioverter defibrillators (CRT-D) may further reduce the risk of sudden cardiac death (SCD). However, economic evaluations have shown

  13. Patient-reported outcomes in Danish implantable cardioverter defibrillator patients with a Sprint Fidelis lead advisory notification

    DEFF Research Database (Denmark)

    Pedersen, Susanne S; Versteeg, Henneke; Nielsen, Jens Cosedis

    2011-01-01

    Few studies have investigated the association between implantable cardioverter defibrillators (ICDs) and lead advisory notifications and patient-reported outcomes (PROs). We examined (i) whether the mode used to inform patients about a device advisory is associated with PROs, and (ii) whether...... patients with a lead subject to a device advisory report poorer PROs than non-advisory controls....

  14. Shock and patient preimplantation type D personality are associated with poor health status in patients with implantable cardioverter-defibrillator

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; Tekle, Fetene B; Hoogwegt, Madelein T

    2012-01-01

    Implantable cardioverter-defibrillator (ICD) shock is a critical event to patients associated with well-being after implantation, although other factors may play an equally important role. We compared the association of shock and the patient's preimplantation personality with health status, using...

  15. Automated image quality assessment for chest CT scans.

    Science.gov (United States)

    Reeves, Anthony P; Xie, Yiting; Liu, Shuang

    2018-02-01

    Medical image quality needs to be maintained at standards sufficient for effective clinical reading. Automated computer analytic methods may be applied to medical images for quality assessment. For chest CT scans in a lung cancer screening context, an automated quality assessment method is presented that characterizes image noise and image intensity calibration. This is achieved by image measurements in three automatically segmented homogeneous regions of the scan: external air, trachea lumen air, and descending aorta blood. Profiles of CT scanner behavior are also computed. The method has been evaluated on both phantom and real low-dose chest CT scans and results show that repeatable noise and calibration measures may be realized by automated computer algorithms. Noise and calibration profiles show relevant differences between different scanners and protocols. Automated image quality assessment may be useful for quality control for lung cancer screening and may enable performance improvements to automated computer analysis methods. © 2017 American Association of Physicists in Medicine.

  16. [External cephalic version].

    Science.gov (United States)

    Navarro-Santana, B; Duarez-Coronado, M; Plaza-Arranz, J

    2016-08-01

    To analyze the rate of successful external cephalic versions in our center and caesarean sections that would be avoided with the use of external cephalic versions. From January 2012 to March 2016 external cephalic versions carried out at our center, which were a total of 52. We collected data about female age, gestational age at the time of the external cephalic version, maternal body mass index (BMI), fetal variety and situation, fetal weight, parity, location of the placenta, amniotic fluid index (ILA), tocolysis, analgesia, and newborn weight at birth, minor adverse effects (dizziness, hypotension and maternal pain) and major adverse effects (tachycardia, bradycardia, decelerations and emergency cesarean section). 45% of the versions were unsuccessful and 55% were successful. The percentage of successful vaginal delivery in versions was 84% (4% were instrumental) and 15% of caesarean sections. With respect to the variables studied, only significant differences in birth weight were found; suggesting that birth weight it is related to the outcome of external cephalic version. Probably we did not find significant differences due to the number of patients studied. For women with breech presentation, we recommend external cephalic version before the expectant management or performing a cesarean section. The external cephalic version increases the proportion of fetuses in cephalic presentation and also decreases the rate of caesarean sections.

  17. Piezosurgery in External Dacryocystorhinostomy.

    Science.gov (United States)

    Czyz, Craig N; Fowler, Amy M; Dutton, Jonathan J; Cahill, Kenneth V; Foster, Jill A; Hill, Robert H; Everman, Kelly R; Nabavi, Cameron B

    Dacryocystorhinostomy (DCR) can be performed via an external or endoscopic approach. The use of ultrasonic or piezosurgery has been well described for endoscopic DCRs but is lacking for external DCRs. This study presents a case series of external DCRs performed using piezosurgery evaluating results and complications. Prospective, consecutive case series of patients undergoing primary external DCR for lacrimal drainage insufficiency. A standard external DCR technique was used using 1 of 2 piezosurgery systems for all bone incision. All patients received silicone intubation to the lacrimal system. Surgical outcome was measured in terms of patient-reported epiphora as follows: 1) complete resolution, 2) improvement >50%, 3) improvement 50% improvement. There were 4 patients (7%) who had <50% improvement. There was 1 (2%) intraoperative complication and 2 (4%) postoperative complications recorded. Piezourgery is a viable modality for performing external DCRs. The lack of surgical complications shows a potential for decreased soft tissues damage. The surgical success rate based on patient-reported epiphora is similar to those published for mechanical external DCRs. This modality may benefit the novice surgeon in the reduction of soft and mucosal tissue damage.

  18. Emotions and health: findings from a randomized clinical trial on psychoeducational nursing to patients with implantable cardioverter defibrillator.

    Science.gov (United States)

    Kikkenborg Berg, Selina; Støier, Louise; Moons, Philip; Zwisler, Ann-Dorthe; Winkel, Per; Ulrich Pedersen, Preben

    2015-01-01

    Serious illness will inevitably lead to a fundamental emotional reaction. Traditionally, in interventional treatment or rehabilitation trials, the psychological status of patients with implantable cardioverter defibrillators has been evaluated with anxiety and depression as outcome measures. In caring for these patients, the aim of nursing is to help patients manage life with complex heart disease. The early detection and management of negative emotional response might prevent the development of pathological conditions such as depression. The aims of this study were to (a) describe the trajectory of primary emotions over time in patients with implantable cardioverter defibrillators and (b) examine the potential effects of psychoeducational nursing on primary emotions. During the inclusion period (October 2007 to November 2009), 196 patients with implantable cardioverter defibrillator were randomized (1:1) to rehabilitation versus usual care. Rehabilitation consisted of a psychoeducational nursing component and an exercise training component. This article concerns phase 1, psychoeducational nursing, guided by a theory of nursing, Rosemary Rizzo Parses Human Becoming Practice Methodologies. The outcome measure is the Emotions and Health Scale. The scale consists of 8 primary emotions: joy, agreeableness, surprise, fear, sadness, disgust, anger, and anticipation. Mean (SD) age was 58 (13) years, and 79% of the participants were men. Significant improvements were found in primary emotional responses over time (P .05). Primary emotions are affected after implantable cardioverter defibrillator implantation. Improvements over time were found. However, it was not possible to detect any effect of a short-term psychoeducational nursing intervention. Evaluating the primary emotions might be a good way for nurses to monitor patients' psychological outcomes because the instrument is sensitive to changes over a short period. Further development of early psychoeducational nursing

  19. Improving treatment plan evaluation with automation

    Science.gov (United States)

    Covington, Elizabeth L.; Chen, Xiaoping; Younge, Kelly C.; Lee, Choonik; Matuszak, Martha M.; Kessler, Marc L.; Keranen, Wayne; Acosta, Eduardo; Dougherty, Ashley M.; Filpansick, Stephanie E.

    2016-01-01

    The goal of this work is to evaluate the effectiveness of Plan‐Checker Tool (PCT) which was created to improve first‐time plan quality, reduce patient delays, increase the efficiency of our electronic workflow, and standardize and automate the physics plan review in the treatment planning system (TPS). PCT uses an application programming interface to check and compare data from the TPS and treatment management system (TMS). PCT includes a comprehensive checklist of automated and manual checks that are documented when performed by the user as part of a plan readiness check for treatment. Prior to and during PCT development, errors identified during the physics review and causes of patient treatment start delays were tracked to prioritize which checks should be automated. Nineteen of 33 checklist items were automated, with data extracted with PCT. There was a 60% reduction in the number of patient delays in the six months after PCT release. PCT was successfully implemented for use on all external beam treatment plans in our clinic. While the number of errors found during the physics check did not decrease, automation of checks increased visibility of errors during the physics check, which led to decreased patient delays. The methods used here can be applied to any TMS and TPS that allows queries of the database. PACS number(s): 87.55.‐x, 87.55.N‐, 87.55.Qr, 87.55.tm, 89.20.Bb PMID:27929478

  20. Automation systems for radioimmunoassay

    International Nuclear Information System (INIS)

    Yamasaki, Paul

    1974-01-01

    The application of automation systems for radioimmunoassay (RIA) was discussed. Automated systems could be useful in the second step, of the four basic processes in the course of RIA, i.e., preparation of sample for reaction. There were two types of instrumentation, a semi-automatic pipete, and a fully automated pipete station, both providing for fast and accurate dispensing of the reagent or for the diluting of sample with reagent. Illustrations of the instruments were shown. (Mukohata, S.)

  1. Laboratory Automation and Middleware.

    Science.gov (United States)

    Riben, Michael

    2015-06-01

    The practice of surgical pathology is under constant pressure to deliver the highest quality of service, reduce errors, increase throughput, and decrease turnaround time while at the same time dealing with an aging workforce, increasing financial constraints, and economic uncertainty. Although not able to implement total laboratory automation, great progress continues to be made in workstation automation in all areas of the pathology laboratory. This report highlights the benefits and challenges of pathology automation, reviews middleware and its use to facilitate automation, and reviews the progress so far in the anatomic pathology laboratory. Copyright © 2015 Elsevier Inc. All rights reserved.

  2. Automated cloning methods.; TOPICAL

    International Nuclear Information System (INIS)

    Collart, F.

    2001-01-01

    Argonne has developed a series of automated protocols to generate bacterial expression clones by using a robotic system designed to be used in procedures associated with molecular biology. The system provides plate storage, temperature control from 4 to 37 C at various locations, and Biomek and Multimek pipetting stations. The automated system consists of a robot that transports sources from the active station on the automation system. Protocols for the automated generation of bacterial expression clones can be grouped into three categories (Figure 1). Fragment generation protocols are initiated on day one of the expression cloning procedure and encompass those protocols involved in generating purified coding region (PCR)

  3. Complacency and Automation Bias in the Use of Imperfect Automation.

    Science.gov (United States)

    Wickens, Christopher D; Clegg, Benjamin A; Vieane, Alex Z; Sebok, Angelia L

    2015-08-01

    We examine the effects of two different kinds of decision-aiding automation errors on human-automation interaction (HAI), occurring at the first failure following repeated exposure to correctly functioning automation. The two errors are incorrect advice, triggering the automation bias, and missing advice, reflecting complacency. Contrasts between analogous automation errors in alerting systems, rather than decision aiding, have revealed that alerting false alarms are more problematic to HAI than alerting misses are. Prior research in decision aiding, although contrasting the two aiding errors (incorrect vs. missing), has confounded error expectancy. Participants performed an environmental process control simulation with and without decision aiding. For those with the aid, automation dependence was created through several trials of perfect aiding performance, and an unexpected automation error was then imposed in which automation was either gone (one group) or wrong (a second group). A control group received no automation support. The correct aid supported faster and more accurate diagnosis and lower workload. The aid failure degraded all three variables, but "automation wrong" had a much greater effect on accuracy, reflecting the automation bias, than did "automation gone," reflecting the impact of complacency. Some complacency was manifested for automation gone, by a longer latency and more modest reduction in accuracy. Automation wrong, creating the automation bias, appears to be a more problematic form of automation error than automation gone, reflecting complacency. Decision-aiding automation should indicate its lower degree of confidence in uncertain environments to avoid the automation bias. © 2015, Human Factors and Ergonomics Society.

  4. Energy policy and externalities

    International Nuclear Information System (INIS)

    Bertel, E.; Fraser, P.

    2002-01-01

    External costs of energy have been assessed in a number of authoritative and reliable studies based upon widely accepted methodologies such as life cycle analysis (LCA). However, although those costs are recognised by most stakeholders and decision makers, results from analytical work on externalities and LCA studies are seldom used in policy making. The International Energy Agency (IEA) and the Nuclear Energy Agency (NEA) convened a joint workshop in November 2001 to offer experts and policy makers an opportunity to present state-of-the-art results from analytical work on externalities and debate issues related to the relevance of external costs and LCA for policy-making purposes. The findings from the workshop highlight the need for further work in the field and the potential rote of international organisations like the IEA and the NEA in this context. (authors)

  5. Device orientation of a leadless pacemaker and subcutaneous implantable cardioverter-defibrillator in canine and human subjects and the effect on intrabody communication

    NARCIS (Netherlands)

    Quast, Anne-Floor B. E.; Tjong, Fleur V. Y.; Koop, Brendan E.; Wilde, Arthur A. M.; Knops, Reinoud E.; Burke, Martin C.

    2018-01-01

    The development of communicating modular cardiac rhythm management systems relies on effective intrabody communication between a subcutaneous implantable cardioverter-defibrillator (S-ICD) and a leadless pacemaker (LP), using conducted communication. Communication success is affected by the LP and

  6. Prophylactic implantable defibrillator in patients with arrhythmogenic right ventricular cardiomyopathy/dysplasia and no prior ventricular fibrillation or sustained ventricular tachycardia.

    LENUS (Irish Health Repository)

    Corrado, Domenico

    2010-09-21

    The role of implantable cardioverter-defibrillator (ICD) in patients with arrhythmogenic right ventricular cardiomyopathy\\/dysplasia and no prior ventricular fibrillation (VF) or sustained ventricular tachycardia is an unsolved issue.

  7. Type D personality is associated with increased anxiety and depressive symptoms in patients with an implantable cardioverter defibrillator and their partners

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; van Domburg, Ron T; Theuns, Dominic A M J

    2014-01-01

    We investigated the prevalence of anxiety and depressive symptoms in patients with an implantable cardioverter defibrillator (ICD) and their partners, and the role of personality factors and social support as determinants of distress....

  8. Automated data collection based on RoboDiff at the ESRF beamline MASSIF-1

    Energy Technology Data Exchange (ETDEWEB)

    Nurizzo, Didier, E-mail: Didier.nurizzo@esrf.fr; Guichard, Nicolas; McSweeney, Sean; Theveneau, Pascal; Guijarro, Matias; Svensson, Olof; Mueller-Dieckmann, Christoph; Leonard, Gordon [ESRF, The European Synchrotron, 71, Avenue des Martyrs,CS 40220, 38043 Grenoble (France); Bowler, Matthew W. [EMBL Grenoble Outstation, 71 Avenue des Martyrs, CS90181, 38042 Grenoble Cedex 9 (France)

    2016-07-27

    The European Synchrotron Radiation Facility has a long standing history in the automation of experiments in Macromolecular Crystallography. MASSIF-1 (Massively Automated Sample Screening and evaluation Integrated Facility), a beamline constructed as part of the ESRF Upgrade Phase I program, has been open to the external user community since July 2014 and offers a unique completely automated data collection service to both academic and industrial structural biologists.

  9. Pacemakers and implantable cardioverter defibrillators, unknown to chest radiography: Review, complications and systematic reading

    International Nuclear Information System (INIS)

    Alandete Germán, Salvador Pascual; Isarria Vidal, Santiago; Domingo Montañana, María Luisa; De la vía Oraá, Esperanza; Vilar Samper, José

    2015-01-01

    Highlights: •Radiologists have an important function in the evaluation of these devices. •We revise their radiological appearances and possible complications. •The knowledge in normal aspects and complications is important for radiologist. •To ensure an accurate reading of the chest x-ray, we present a systematic approach. -- Abstract: Chest X-ray is the imaging technique of choice for an initial study of pacemakers and implantable cardio-defibrillators (ICD). Radiologists have an important role in the evaluation of its initial placement and in the assessment during its follow-up. For this reason, it is necessary to know not only the different existing devices and its components but also the reasons of malfunction or possible complications. The purpose of this article is to do a systematic review of the different types of pacemakers and ICD. We review their usual radiological appearances, the possible complications which might take place and its causes of malfunctioning

  10. Cardiac Resynchronization Therapy Defibrillator Treatment in a Child with Heart Failure and Ventricular Arrhythmia

    Directory of Open Access Journals (Sweden)

    Hak Ju Kim

    2016-08-01

    Full Text Available Cardiac resynchronization therapy (CRT is a new treatment for refractory heart failure. However, most patients with heart failure treated with CRT are adults, middle-aged or older with idiopathic or ischemic dilated cardiomyopathy. We treated a 12-year-old boy, who was transferred after cardiac arrest, with dilated cardiomyopathy, left bundle-branch block, and ventricular tachycardia. We performed cardiac resynchronization therapy with a defibrillator (CRT-D. After CRT-D, left ventricular ejection fraction improved from 22% to 4 4% a ssessed by e chocardiog ram 1 year p ostoperatively. On e lectrocardiog ram, Q RS d uration was shortened from 206 to 144 ms. The patient’s clinical symptoms also improved. For pediatric patients with refractory heart failure and ventricular arrhythmia, CRT-D could be indicated as an effective therapeutic option.

  11. Relation of statin therapy to psychological functioning in patients with an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Hoogwegt, Madelein T; Theuns, Dominic A M J; Kupper, Nina

    2013-01-01

    Statin therapy is an important secondary prevention measure in cardiovascular disease. However, the side effects associated with statin use could potentially affect patients' quality of life. Little is known about the influence of statin therapy on the well-being and health status of cardiac...... patients, in general, and patients with an implantable cardioverter defibrillator (ICD), in particular. We investigated the association between statin therapy and symptoms of anxiety and depression and patients' health status during the 12 months after implantation, reckoning with statin type and dosage...... of statin type, dosage, and other potential confounders. The associations between statin therapy and depression (p = 0.06) and statin therapy and physical functioning (p = 0.05) were borderline significant, and no association was found with anxiety (p >0.05). In conclusion, statin therapy was associated...

  12. The Subcutaneous Implantable Cardioverter-Defibrillator: A Practical Review and Real-World Use and Application

    Directory of Open Access Journals (Sweden)

    Mark E. Panna Jr, MD, FACC, FHRS

    2016-02-01

    Full Text Available The subcutaneous implantable cardioverter-defibrillator (ICD is a novel technology using a subcutaneous (extrathoracic system for treatment of potential lethal ventricular arrhythmias. It avoids many of the risks of transvenous ICD implantation. It may be considered in patients having an ICD indication who do not have a pacing and/or cardiac resynchronization therapy indication, and who are unlikely to benefit from antitachycardia pacing therapy. We review patient selection, system components, the implantation technique, and screening considerations for subcutaneous ICD implantation. Its uses in specific patient populations, including children, patients with congenital heart disease, hypertrophic cardiomyopathy, or end-stage renal disease, and patients with preexisting pacemakers, are highlighted. Areas of future investigation are reviewed, including potential use with leadless pacing and magnetic resonance imaging.

  13. Echocardiography in patients with complications related to pacemakers and cardiac defibrillators.

    Science.gov (United States)

    Almomani, Ahmed; Siddiqui, Khadija; Ahmad, Masood

    2014-03-01

    The evolving indications and uses for implantable cardiac devices have led to a significant increase in the number of implanted devices each year. Implantation of endocardial leads for permanent pacemakers and cardiac defibrillators can cause many delayed complications. Complications may be mechanical and related to the interaction of the device leads with the valves and endomyocardium, e.g., perforation, infection, and thrombosis, or due to the electrical pacing of the myocardium and conduction abnormalities, e.g., left ventricular dyssynchrony. Tricuspid regurgitation, another delayed complication in these patients, may be secondary to both mechanical and pacing effects of the device leads. Echocardiography plays an important role in the diagnosis of these device-related complications. Both two-dimensional transthoracic echocardiography and transesophageal echocardiography provide useful diagnostic information. Real time three-dimensional echocardiography is a novel technique that can further enhance the detection of lead-related complications. © 2013, Wiley Periodicals, Inc.

  14. Arrhythmogenic Right Ventricular Cardiomyopathy: Risk Stratification and Indications for Defibrillator Therapy.

    Science.gov (United States)

    Zorzi, Alessandro; Rigato, Ilaria; Bauce, Barbara; Pilichou, Kalliopi; Basso, Cristina; Thiene, Gaetano; Iliceto, Sabino; Corrado, Domenico

    2016-06-01

    Arrhythmogenic right ventricular cardiomyopathy (ARVC) is a genetically determined disease which predisposes to life-threatening ventricular arrhythmias. The main goal of ARVC therapy is prevention of sudden cardiac death (SCD). Implantable cardioverter defibrillator (ICD) is the most effective therapy for interruption of potentially lethal ventricular tachyarrhythmias. Despite its life-saving potential, ICD implantation is associated with a high rate of complications and significant impact on quality of life. Accurate risk stratification is needed to identify individuals who most benefit from the therapy. While there is general agreement that patients with a history of cardiac arrest or hemodynamically unstable ventricular tachycardia are at high risk of SCD and needs an ICD, indications for primary prevention remain a matter of debate. The article reviews the available scientific evidence and guidelines that may help to stratify the arrhythmic risk of ARVC patients and guide ICD implantation. Other therapeutic strategies, either alternative or additional to ICD, will be also addressed.

  15. Trajectories of Patient-Reported Health Status in Patients With an Implantable Cardioverter Defibrillator

    DEFF Research Database (Denmark)

    Mastenbroek, Mirjam H; Denollet, Johan; Versteeg, Henneke

    2015-01-01

    , no use of ACE inhibitors, psychotropic medication, negative affectivity, and type D personality were identified as independent determinants of poorer mental health status. In conclusion, the population with an ICD seems to be heterogeneous in terms of patient-reported physical and mental health status......To date, no study has assessed the course of patient-reported health status in patients with an implantable cardioverter defibrillator (ICD). Studying health status trajectories and their baseline determinants would permit the identification of patients at risk for poor health outcomes after ICD...... implantation. A combined cohort of 1,222 patients with an ICD (79% men; age = 61.4 [11.2] years) completed the 12-Item Short-Form Health Survey at baseline and 2 to 3 months and 12 to 14 months after implantation. Latent class analyses were used to identify trajectories and predictors of health status over...

  16. Interactions between pacing and arrhythmia detection algorithms in the dual chamber implantable cardioverter defibrillator.

    Science.gov (United States)

    Dijkman, B; Wellens, H J

    2001-09-01

    Dual chamber implantable cardioverter defibrillator (ICD) combines the possibility to detect and treat ventricular and atrial arrhythmias with the possibility of modern heart stimulation techniques. Advanced pacing algorithms together with extended arrhythmia detection capabilities can give rise to new types of device-device interactions. Some of the possible interactions are illustrated by four cases documented in four models of dual chamber ICDs. Functioning of new features in dual chamber devices is influenced by the fact that the pacemaker is not a separate device but a part of the ICD system and that both are being used in a patient with arrhythmia. Programming measures are suggested to optimize use of new pacing algorithms while maintaining correct arrhythmia detection.

  17. Attitudes towards implantable cardioverter-defibrillator therapy: a national survey in Danish health-care professionals

    DEFF Research Database (Denmark)

    Johansen, Jens B; Mortensen, Peter T; Videbæk, Regitze

    2011-01-01

    Aims The aim of this study was to examine health-care professionals attitudes towards implantable cardioverter-defibrillator (ICD) therapy and issues discussed with patients. Methods and results Survey of 209 health-care professionals providing specialized treatment and care of ICD patients......-physicians. Physicians were less likely to believe that their personal attitude towards ICD treatment has no influence on how they deal professionally with patients (27.8 vs. 43.6%; P = 0.04). Physicians and non-physicians were equally positive towards ICD therapy as primary prophylaxis in ischaemic cardiomyopathy (87...... discussing ICD treatment with candidate patients. At the same time, physicians are more aware that their attitude towards ICD treatment may influence how they deal professionally with patients compared with non-physicians....

  18. Illness perceptions in patients with heart failure and an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Timmermans, I.; Versteeg, H.; Meine, Mathias M

    2017-01-01

    Background Patients' illness perceptions are associated with psychological wellbeing and can be measured with the Brief Illness Perception Questionnaire (B-IPQ). However, little is known about illness perceptions in patients with heart failure. We examined the dimensional structure, validity...... and clinical and psychological correlates of the B-IPQ in Dutch, French and German patients with heart failure and an implantable cardioverter defibrillator (ICD). Method European heart failure patients (n=585) participating in the REMOTE-CIED study completed a set of questionnaires 1–2weeks post ICD.......69, with the ‘Consequences’ subscale being more internally consistent (α=0.80). Both the B-IPQ and its ‘Consequences’ subscale were significantly correlated with a number of psychological characteristics, but not with clinical characteristics. Multivariable logistic regression analysis indicated that threatening illness...

  19. Patients' perspective on deactivation of the implantable cardioverter-defibrillator near the end of life

    DEFF Research Database (Denmark)

    Pedersen, Susanne S.; Chaitsing, Rismy; Szili-Torok, Tamas

    2013-01-01

    (67%) completed the survey. Most patients (68%) were aware that it is possible to turn the ICD off, and 95% believed it is important to inform patients about the possibility. Of the patients completing the survey, 84% indicated a choice for or against deactivation. Psychological morbidity......Recent guidelines have emphasized the importance of discussing the issue of deactivation near the end of life with patients with an implantable cardioverter-defibrillator (ICD). Few studies have examined the patient perspective and patients' wishes. We examined patients' knowledge and wishes...... for information; and the prevalence and correlates of a favorable attitude toward deactivation. Three cohorts of ICD patients (n = 440) extracted from our institutional database were asked to complete a survey that included a vignette about deactivation near the end of life. Of the 440 patients approached, 294...

  20. Relation between emotional distress and heart rate variability in patients with an implantable cardioverter-defibrillator

    DEFF Research Database (Denmark)

    Hoogwegt, Madelein T; Pedersen, Susanne S.; Theuns, Dominic A M J

    2014-01-01

    We investigated the relationship between Type D personality, depression, and anxiety, and heart rate variability (HRV) in 64 patients with an implantable cardioverter-defibrillator (ICD). HRV was obtained via 24-h Holter monitoring, and 24-h, 30-min daytime rest and 30-min nighttime sleep HRV were...... analyzed. In adjusted analyses, significant associations (standard deviation of normal-to-normal [NN] intervals [SDNN]: p = .043; standard deviation of NN intervals over 5-min periods [SDANN]: p = .010) and a trend (HRV triangular index: p = .09) were found for Type D personality, and trends were found...... = .043). A Benjamini-Hochberg correction for multiple testing led to reduction of the number of significant relationships, but there was still support for lower autonomic control patients with Type D personality and depression. Future research with larger sample sizes is warranted....

  1. Gender disparities in psychological distress and quality of life among patients with an implantable cardioverter defibrillator

    DEFF Research Database (Denmark)

    Brouwers, Corline; van den Broek, Krista C; Denollet, Johan

    2011-01-01

    A subset of patients with an implantable cardioverter defibrillator (ICD) reports psychological distress and poor quality of life (QoL). Gender is one of the factors that has been proposed to explain individual differences in these outcomes. In this viewpoint, we (1) review the evidence for gender...... disparities in psychological distress and QoL in ICD patients by means of a systematic review, and (2) provide recommendations for future research and clinical implications. A systematic search of the literature identified 18 studies with a sample size ≥ 100 that examined gender disparities in anxiety....../depression and QoL in ICD patients (mean prevalence of women = 21%; mean age = 62 years). Our review shows that there is insufficient evidence to conclude that gender per se is a major autonomous predictor for disparities in psychological distress and QoL in ICD patients. Women had a higher prevalence of anxiety...

  2. Automated System Marketplace 1994.

    Science.gov (United States)

    Griffiths, Jose-Marie; Kertis, Kimberly

    1994-01-01

    Reports results of the 1994 Automated System Marketplace survey based on responses from 60 vendors. Highlights include changes in the library automation marketplace; estimated library systems revenues; minicomputer and microcomputer-based systems; marketplace trends; global markets and mergers; research needs; new purchase processes; and profiles…

  3. Automation in Warehouse Development

    NARCIS (Netherlands)

    Hamberg, R.; Verriet, J.

    2012-01-01

    The warehouses of the future will come in a variety of forms, but with a few common ingredients. Firstly, human operational handling of items in warehouses is increasingly being replaced by automated item handling. Extended warehouse automation counteracts the scarcity of human operators and

  4. Order Division Automated System.

    Science.gov (United States)

    Kniemeyer, Justin M.; And Others

    This publication was prepared by the Order Division Automation Project staff to fulfill the Library of Congress' requirement to document all automation efforts. The report was originally intended for internal use only and not for distribution outside the Library. It is now felt that the library community at-large may have an interest in the…

  5. Automate functional testing

    Directory of Open Access Journals (Sweden)

    Ramesh Kalindri

    2014-06-01

    Full Text Available Currently, software engineers are increasingly turning to the option of automating functional tests, but not always have successful in this endeavor. Reasons range from low planning until over cost in the process. Some principles that can guide teams in automating these tests are described in this article.

  6. Automation and robotics

    Science.gov (United States)

    Montemerlo, Melvin

    1988-01-01

    The Autonomous Systems focus on the automation of control systems for the Space Station and mission operations. Telerobotics focuses on automation for in-space servicing, assembly, and repair. The Autonomous Systems and Telerobotics each have a planned sequence of integrated demonstrations showing the evolutionary advance of the state-of-the-art. Progress is briefly described for each area of concern.

  7. Automating the Small Library.

    Science.gov (United States)

    Skapura, Robert

    1987-01-01

    Discusses the use of microcomputers for automating school libraries, both for entire systems and for specific library tasks. Highlights include available library management software, newsletters that evaluate software, constructing an evaluation matrix, steps to consider in library automation, and a brief discussion of computerized card catalogs.…

  8. Magnetic resonance imaging at 1.5-T in patients with implantable cardioverter-defibrillators.

    Science.gov (United States)

    Naehle, Claas P; Strach, Katharina; Thomas, Daniel; Meyer, Carsten; Linhart, Markus; Bitaraf, Sascha; Litt, Harold; Schwab, Jörg Otto; Schild, Hans; Sommer, Torsten

    2009-08-04

    Our aim was to establish and evaluate a strategy for safe performance of magnetic resonance imaging (MRI) at 1.5-T in patients with implantable cardioverter-defibrillators (ICDs). Expanding indications for ICD placement and MRI becoming the imaging modality of choice for many indications has created a growing demand for MRI in ICD patients, which is still considered an absolute contraindication. Non-pacemaker-dependent ICD patients with a clinical need for MRI were included in the study. To minimize radiofrequency-related lead heating, the specific absorption rate was limited to 2 W/kg. ICDs were reprogrammed pre-MRI to avoid competitive pacing and potential pro-arrhythmia: 1) the lower rate limit was programmed as low as reasonably achievable; and 2) arrhythmia detection was programmed on, but therapy delivery was programmed off. Patients were monitored using electrocardiography and pulse oximetry. All ICDs were interrogated before and after the MRI examination and after 3 months, including measurement of pacing capture threshold, lead impedance, battery voltage, and serum troponin I. Eighteen ICD patients underwent a total of 18 MRI examinations at 1.5-T; all examinations were completed safely. All ICDs could be interrogated and reprogrammed normally post-MRI. No significant changes of pacing capture threshold, lead impedance, and serum troponin I were observed. Battery voltage decreased significantly from pre- to post-MRI. In 2 MRI examinations, oversensing of radiofrequency noise as ventricular fibrillation occurred. However, no attempt at therapy delivery was made. MRI of non-pacemaker-dependent ICD patients can be performed with an acceptable risk/benefit ratio under controlled conditions by taking both MRI- and pacemaker-related precautions. (Implantable Cardioverter Defibrillators and Magnetic Resonance Imaging of the Heart at 1.5-Tesla; NCT00356239).

  9. Electrical storm in patients with implantable cardioverter-defibrillators: can it be forecast?

    Science.gov (United States)

    Emkanjoo, Zahra; Alihasani, Narges; Alizadeh, Abolfath; Tayyebi, Mohammad; Bonakdar, Hamid; Barakpour, Hamid; Sadr-Ameli, Mohammad Ali

    2009-01-01

    The aim of this retrospective study was to determine the prevalence and predictors of electrical storm in 227 patients who had received implantable cardioverter-defibrillators (ICDs) and had been monitored for 31.7 +/- 15.6 months. Of these, 174 (77%) were men. The mean age was 55.8 +/- 15.5 years (range, 20-85 yr), and the mean left ventricular ejection fraction (LVEF) was 0.30 +/- 0.14. One hundred forty-six of the patients (64%) had underlying coronary artery disease. Cardioverter-defibrillators were implanted for secondary (80%) and primary (20%) prevention. Of the 227 patients, 117 (52%) experienced events that required ICD therapy. Thirty patients (mean age, 57.26 +/- 14.3 yr) had > or = 3 episodes requiring ICD therapy in a 24-hour period and were considered to have electrical storm. The mean number of events was 12.75 +/- 15 per patient. Arrhythmia-clustering occurred an average of 6.1 +/- 6.7 months after ICD implantation. Clinical variables with the most significant association with electrical storm were low LVEF (P = 0.04; hazard ratio of 0.261, and 95% confidence interval of 0.08-0.86) and higher use of class IA antiarrhythmic drugs (P = 0.018, hazard ratio of 3.84, and 95% confidence interval of 1.47-10.05). Amiodarone treatment and use of beta-blockers were not significant predictors when subjected to multivariate analysis. We conclude that electrical storm is most likely to occur in patients with lower LVEF and that the use of Class IA antiarrhythmic drugs is a risk factor.

  10. Electrical storm presages nonsudden death: the antiarrhythmics versus implantable defibrillators (AVID) trial.

    Science.gov (United States)

    Exner, D V; Pinski, S L; Wyse, D G; Renfroe, E G; Follmann, D; Gold, M; Beckman, K J; Coromilas, J; Lancaster, S; Hallstrom, A P

    2001-04-24

    Electrical storm, multiple temporally related episodes of ventricular tachycardia (VT) or ventricular fibrillation (VF), is a frequent problem among recipients of implantable cardioverter defibrillators (ICDs). However, insufficient data exist regarding its prognostic significance. This analysis includes 457 patients who received an ICD in the Antiarrhythmics Versus Implantable Defibrillators (AVID) trial and who were followed for 31 +/- 13 months. Electrical storm was defined as > or = 3 separate episodes of VT/VF within 24 hours. Characteristics and survival of patients surviving electrical storm (n = 90), those with VT/VF unrelated to electrical storm (n = 184), and the remaining patients (n = 183) were compared. The 3 groups differed in terms of ejection fraction, index arrhythmia, revascularization status, and baseline medication use. Survival was evaluated using time-dependent Cox modeling. Electrical storm occurred 9.2 +/- 11.5 months after ICD implantation, and most episodes (86%) were due to VT. Electrical storm was a significant risk factor for subsequent death, independent of ejection fraction and other prognostic variables (relative risk [RR], 2.4; 95% confidence interval [CI], 1.3 to 4.2; P = 0.003), but VT/VF unrelated to electrical storm was not (RR, 1.0; 95% CI, 0.6 to 1.7; P = 0.9). The risk of death was greatest 3 months after electrical storm (RR, 5.4; 95% Cl, 2.4 to 12.3; P = 0.0001) and diminished beyond this time (RR, 1.9; 95% CI, 1.0 to 3.6; P=0.04). Electrical storm is an important, independent marker for subsequent death among ICD recipients, particularly in the first 3 months after its occurrence. However, the development of VT/VF unrelated to electrical storm does not seem to be associated with an increased risk of subsequent death.

  11. Outcomes with single-coil versus dual-coil implantable cardioverter defibrillators: a meta-analysis.

    Science.gov (United States)

    Sunderland, Nicholas; Kaura, Amit; Murgatroyd, Francis; Dhillon, Para; Scott, Paul A

    2018-03-01

    Dual-coil implantable cardioverter defibrillator (ICD) leads have traditionally been used over single-coil leads due to concerns regarding high defibrillation thresholds (DFT) and consequent poor shock efficacy. However, accumulating evidence suggests that this position may be unfounded and that dual-coil leads may also be associated with higher complication rates during lead extraction. This meta-analysis collates data comparing dual- and single-coil ICD leads. Electronic databases were systematically searched for randomized controlled trials (RCT) and non-randomized studies comparing single-coil and dual-coil leads. The mean differences in DFT and summary estimates of the odds-ratio (OR) for first-shock efficacy and the hazard-ratio (HR) for all-cause mortality were calculated using random effects models. Eighteen studies including a total of 138,124 patients were identified. Dual-coil leads were associated with a lower DFT compared to single coil leads (mean difference -0.83J; 95% confidence interval [CI] -1.39--0.27; P = 0.004). There was no difference in the first-shock success rate with dual-coil compared to single-coil leads (OR 0.74; 95%CI 0.45-1.21; P=0.22). There was a significantly lower risk of all-cause mortality associated with single-coil leads (HR 0.91; 95%CI 0.86-0.95; P dual-coil leads. The mortality benefit with single-coil leads most likely represents patient selection bias. Given the increased risk and complexity of extracting dual-coil leads, centres should strongly consider single-coil ICD leads as the lead of choice for routine new left-sided ICD implants. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2017. For permissions, please email: journals.permissions@oup.com.

  12. Physicians’ knowledge and attitudes in Saudi Arabia regarding implantable cardiac defibrillators

    Directory of Open Access Journals (Sweden)

    Tariq Alhogbani

    2018-04-01

    Full Text Available Objectives: To evaluate knowledge and attitude of physicians involved in the management of patients with heart failure regarding implantable cardioverter-defibrillator (ICD. Methods: We conducted personal interviews with physicians involved in treating patients with heart failure. Between October 2015 and February 2016, the study was conducted in hospitals in the Riyadh region where no cardiac electrophysiology service was available. Every participant was met in person and received an oral questionnaire that aimed to assess basic knowledge regarding ICD indications and benefits. Results: Sixty-three physicians were met from 13 hospitals (14 consultants and 49 specialists. Forty-one percent of participants use the recommended cut-off level of left ventricular ejection fraction (LVEF which is ≤35% as the LVEF criterion for ICD referral in patients with cardiomyopathy. Only 50% of the consultants use ≤35% as the LVEF criterion for ICD referral. Seventy percent of the participants thought that ICD may improve heart failure symptoms. Forty-eight percent of physicians have a defined channel to refer patients to higher centers for ICD implant. There was no statistically significant difference between physicians’ knowledge when we categorized them according to three different factors: (1 physician’s specialty (cardiology vs. internal medicine; (2 physician’s degree (consultant vs. specialist; and (3 physician’s location (inside vs. outside Riyadh city. Conclusion: There is a lack of knowledge of current clinical guidelines regarding ICD implantation for patients with heart failure at general hospitals in Saudi Arabia. This finding highlights the need to improve the dissemination of guidelines to practitioners involved in managing patients with heart failure in an effort to improve ICD utilization. Keywords: Cardiac defibrillator, Heart failure, Physicians’ knowledge, Saudi Arabia

  13. Automated model building

    CERN Document Server

    Caferra, Ricardo; Peltier, Nicholas

    2004-01-01

    This is the first book on automated model building, a discipline of automated deduction that is of growing importance Although models and their construction are important per se, automated model building has appeared as a natural enrichment of automated deduction, especially in the attempt to capture the human way of reasoning The book provides an historical overview of the field of automated deduction, and presents the foundations of different existing approaches to model construction, in particular those developed by the authors Finite and infinite model building techniques are presented The main emphasis is on calculi-based methods, and relevant practical results are provided The book is of interest to researchers and graduate students in computer science, computational logic and artificial intelligence It can also be used as a textbook in advanced undergraduate courses

  14. Automation in Immunohematology

    Directory of Open Access Journals (Sweden)

    Meenu Bajpai

    2012-01-01

    Full Text Available There have been rapid technological advances in blood banking in South Asian region over the past decade with an increasing emphasis on quality and safety of blood products. The conventional test tube technique has given way to newer techniques such as column agglutination technique, solid phase red cell adherence assay, and erythrocyte-magnetized technique. These new technologies are adaptable to automation and major manufacturers in this field have come up with semi and fully automated equipments for immunohematology tests in the blood bank. Automation improves the objectivity and reproducibility of tests. It reduces human errors in patient identification and transcription errors. Documentation and traceability of tests, reagents and processes and archiving of results is another major advantage of automation. Shifting from manual methods to automation is a major undertaking for any transfusion service to provide quality patient care with lesser turnaround time for their ever increasing workload. This article discusses the various issues involved in the process.

  15. Automation in Warehouse Development

    CERN Document Server

    Verriet, Jacques

    2012-01-01

    The warehouses of the future will come in a variety of forms, but with a few common ingredients. Firstly, human operational handling of items in warehouses is increasingly being replaced by automated item handling. Extended warehouse automation counteracts the scarcity of human operators and supports the quality of picking processes. Secondly, the development of models to simulate and analyse warehouse designs and their components facilitates the challenging task of developing warehouses that take into account each customer’s individual requirements and logistic processes. Automation in Warehouse Development addresses both types of automation from the innovative perspective of applied science. In particular, it describes the outcomes of the Falcon project, a joint endeavour by a consortium of industrial and academic partners. The results include a model-based approach to automate warehouse control design, analysis models for warehouse design, concepts for robotic item handling and computer vision, and auton...

  16. Malignant external otitis

    International Nuclear Information System (INIS)

    Dupuch, K.M.; Iryboz, T.; Firat, M.; Levy, C.; Tubiana, J.M.

    1991-01-01

    This paper illustrates the value of CT and MR in early diagnosis and spread of malignant external otitis. The authors retrospectively analyzed 15 patients with proved malignant external otitis examined with postcontrast high-resolution CT (15/15) and MR (6/15) (T1- and T2-weighting). Gallium studies were done in 6/15 patients. Early diagnosis was made when CT demonstrated a soft-tissue mass of the external auditory canal associated with scattered zones of cortical bone erosions (13/15). Spread of the disease was better delineated by MR than CT, especially skull base extension (6/15). Temporomandibular joint involvement with extension into parotid or/and masticator spaces 6/15 was as well detected with CT as with MR. If CT remains the first and best procedure for diagnosis, MR - despite its cost - appears a good procedure to depict exact anatomic spread, allowing therapeutic management

  17. Productivity Change and Externalities

    DEFF Research Database (Denmark)

    Kravtsova, Victoria

    2014-01-01

    This paper contributes to the analysis of the impact of externalities on the host country's total factor productivity by taking into account different dimensions of spillover effects. Namely, engagement in exporting and foreign ownership is generally perceived as being beneficial to individual...... firms and the economy as a whole. The approach used in the current research accounts for different internal as well as external factors that individual firms face and evaluates the effect on changes in productivity, technology as well as the efficiency of domestic firms. The empirical analysis focuses...... on Hungary. While the country leads the group of post-socialist countries in the amount of attracted foreign direct investments (FDI) the effect of this policy on the economy remains unclear. The research finds that different externalities play a different role in productivity, technological and efficiency...

  18. Externality or sustainability economics?

    International Nuclear Information System (INIS)

    Bergh, Jeroen C.J.M. van den

    2010-01-01

    In an effort to develop 'sustainability economics' Baumgaertner and Quaas (2010) neglect the central concept of environmental economics-'environmental externality'. This note proposes a possible connection between the concepts of environmental externality and sustainability. In addition, attention is asked for other aspects of 'sustainability economics', namely the distinction weak/strong sustainability, spatial sustainability and sustainable trade, distinctive sustainability policy, and the ideas of early 'sustainability economists'. I argue that both sustainability and externalities reflect a systems perspective and propose that effective sustainability solutions require that more attention is given to system feedbacks, notably other-regarding preferences and social interactions, and energy and environmental rebound. The case of climate change and policy is used to illustrate particular statements. As a conclusion, a list of 20 insights and suggestions for research is offered. (author)

  19. Metasurface external cavity laser

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Luyao, E-mail: luyaoxu.ee@ucla.edu; Curwen, Christopher A.; Williams, Benjamin S. [Department of Electrical Engineering, University of California, Los Angeles, California 90095 (United States); California NanoSystems Institute, University of California, Los Angeles, California 90095 (United States); Hon, Philip W. C.; Itoh, Tatsuo [Department of Electrical Engineering, University of California, Los Angeles, California 90095 (United States); Chen, Qi-Sheng [Northrop Grumman Aerospace Systems, Redondo Beach, California 90278 (United States)

    2015-11-30

    A vertical-external-cavity surface-emitting-laser is demonstrated in the terahertz range, which is based upon an amplifying metasurface reflector composed of a sub-wavelength array of antenna-coupled quantum-cascade sub-cavities. Lasing is possible when the metasurface reflector is placed into a low-loss external cavity such that the external cavity—not the sub-cavities—determines the beam properties. A near-Gaussian beam of 4.3° × 5.1° divergence is observed and an output power level >5 mW is achieved. The polarized response of the metasurface allows the use of a wire-grid polarizer as an output coupler that is continuously tunable.

  20. Factors to Consider When Implementing Automated Software Testing

    Science.gov (United States)

    2016-11-10

    development and integration is a continuous process throughout the acquisition life cycle . Automated Software Testing can improve testing capabilities...requires a lab, conference room, or both, and whether it should be located in- house or an external facility. 2. Ensure space is adequate to support team

  1. Automated computation of one-loop integrals in massless theories

    International Nuclear Information System (INIS)

    Hameren, A. van; Vollinga, J.; Weinzierl, S.

    2005-01-01

    We consider one-loop tensor and scalar integrals, which occur in a massless quantum field theory, and we report on the implementation into a numerical program of an algorithm for the automated computation of these one-loop integrals. The number of external legs of the loop integrals is not restricted. All calculations are done within dimensional regularization. (orig.)

  2. The External Mind

    DEFF Research Database (Denmark)

    , Extended Mind and Distributed Cognition by Claudio Paolucci pp. 69-96 The Social Horizon of Embodied Language and Material Symbols by Riccardo Fusaroli pp. 97-123 Semiotics and Theories of Situated/Distributed Action and Cognition: a Dialogue and Many Intersections by Tommaso Granelli pp. 125-167 Building......The External Mind: an Introduction by Riccardo Fusaroli, Claudio Paolucci pp. 3-31 The sign of the Hand: Symbolic Practices and the Extended Mind by Massimiliano Cappuccio, Michael Wheeler pp. 33-55 The Overextended Mind by Shaun Gallagher pp. 57-68 The "External Mind": Semiotics, Pragmatism...

  3. Systematic review automation technologies

    Science.gov (United States)

    2014-01-01

    Systematic reviews, a cornerstone of evidence-based medicine, are not produced quickly enough to support clinical practice. The cost of production, availability of the requisite expertise and timeliness are often quoted as major contributors for the delay. This detailed survey of the state of the art of information systems designed to support or automate individual tasks in the systematic review, and in particular systematic reviews of randomized controlled clinical trials, reveals trends that see the convergence of several parallel research projects. We surveyed literature describing informatics systems that support or automate the processes of systematic review or each of the tasks of the systematic review. Several projects focus on automating, simplifying and/or streamlining specific tasks of the systematic review. Some tasks are already fully automated while others are still largely manual. In this review, we describe each task and the effect that its automation would have on the entire systematic review process, summarize the existing information system support for each task, and highlight where further research is needed for realizing automation for the task. Integration of the systems that automate systematic review tasks may lead to a revised systematic review workflow. We envisage the optimized workflow will lead to system in which each systematic review is described as a computer program that automatically retrieves relevant trials, appraises them, extracts and synthesizes data, evaluates the risk of bias, performs meta-analysis calculations, and produces a report in real time. PMID:25005128

  4. Externalities - an analysis using the EU ExternE-results

    International Nuclear Information System (INIS)

    2003-10-01

    The EU project ExternE quantified the externalities for the different energy technologies. In this work, the ExternE results are used in a MARKAL-analysis for the Nordic countries. The analysis does not go into detail, but gives some interesting indications: The external costs are not fully covered in the Nordic energy systems, the present taxes and charges are not high enough. The emissions from the energy systems would be strongly reduced, if taxes/environmental charges were set at the level ExternE calculate. The emissions from power production would be reduced most. Renewable energy sources and natural gas dominate the energy systems in the ExternE case

  5. Operational proof of automation

    International Nuclear Information System (INIS)

    Jaerschky, R.; Reifenhaeuser, R.; Schlicht, K.

    1976-01-01

    Automation of the power plant process may imply quite a number of problems. The automation of dynamic operations requires complicated programmes often interfering in several branched areas. This reduces clarity for the operating and maintenance staff, whilst increasing the possibilities of errors. The synthesis and the organization of standardized equipment have proved very successful. The possibilities offered by this kind of automation for improving the operation of power plants will only sufficiently and correctly be turned to profit, however, if the application of these technics of equipment is further improved and if its volume is tallied with a definite etc. (orig.) [de

  6. Automation of radioimmunoassay

    International Nuclear Information System (INIS)

    Yamaguchi, Chisato; Yamada, Hideo; Iio, Masahiro

    1974-01-01

    Automation systems for measuring Australian antigen by radioimmunoassay under development were discussed. Samples were processed as follows: blood serum being dispensed by automated sampler to the test tube, and then incubated under controlled time and temperature; first counting being omitted; labelled antibody being dispensed to the serum after washing; samples being incubated and then centrifuged; radioactivities in the precipitate being counted by auto-well counter; measurements being tabulated by automated typewriter. Not only well-type counter but also position counter was studied. (Kanao, N.)

  7. Automated electron microprobe

    International Nuclear Information System (INIS)

    Thompson, K.A.; Walker, L.R.

    1986-01-01

    The Plant Laboratory at the Oak Ridge Y-12 Plant has recently obtained a Cameca MBX electron microprobe with a Tracor Northern TN5500 automation system. This allows full stage and spectrometer automation and digital beam control. The capabilities of the system include qualitative and quantitative elemental microanalysis for all elements above and including boron in atomic number, high- and low-magnification imaging and processing, elemental mapping and enhancement, and particle size, shape, and composition analyses. Very low magnification, quantitative elemental mapping using stage control (which is of particular interest) has been accomplished along with automated size, shape, and composition analysis over a large relative area

  8. Chef infrastructure automation cookbook

    CERN Document Server

    Marschall, Matthias

    2013-01-01

    Chef Infrastructure Automation Cookbook contains practical recipes on everything you will need to automate your infrastructure using Chef. The book is packed with illustrated code examples to automate your server and cloud infrastructure.The book first shows you the simplest way to achieve a certain task. Then it explains every step in detail, so that you can build your knowledge about how things work. Eventually, the book shows you additional things to consider for each approach. That way, you can learn step-by-step and build profound knowledge on how to go about your configuration management

  9. Managing laboratory automation.

    Science.gov (United States)

    Saboe, T J

    1995-01-01

    This paper discusses the process of managing automated systems through their life cycles within the quality-control (QC) laboratory environment. The focus is on the process of directing and managing the evolving automation of a laboratory; system examples are given. The author shows how both task and data systems have evolved, and how they interrelate. A BIG picture, or continuum view, is presented and some of the reasons for success or failure of the various examples cited are explored. Finally, some comments on future automation need are discussed.

  10. Automated PCB Inspection System

    Directory of Open Access Journals (Sweden)

    Syed Usama BUKHARI

    2017-05-01

    Full Text Available Development of an automated PCB inspection system as per the need of industry is a challenging task. In this paper a case study is presented, to exhibit, a proposed system for an immigration process of a manual PCB inspection system to an automated PCB inspection system, with a minimal intervention on the existing production flow, for a leading automotive manufacturing company. A detailed design of the system, based on computer vision followed by testing and analysis was proposed, in order to aid the manufacturer in the process of automation.

  11. Operational proof of automation

    International Nuclear Information System (INIS)

    Jaerschky, R.; Schlicht, K.

    1977-01-01

    Automation of the power plant process may imply quite a number of problems. The automation of dynamic operations requires complicated programmes often interfering in several branched areas. This reduces clarity for the operating and maintenance staff, whilst increasing the possibilities of errors. The synthesis and the organization of standardized equipment have proved very successful. The possibilities offered by this kind of automation for improving the operation of power plants will only sufficiently and correctly be turned to profit, however, if the application of these equipment techniques is further improved and if it stands in a certain ratio with a definite efficiency. (orig.) [de

  12. Using left-ventricular-only pacing to eliminate T-wave oversensing in a biventricular implantable cardiac defibrillator.

    Science.gov (United States)

    Khoo, Clarence; Bennett, Matthew; Chakrabarti, Santabhanu; LeMaitre, John; Tung, Stanley K K

    2013-02-01

    A man aged 75 years and with nonischemic cardiomyopathy had implantation of a biventricular implantable cardiac defibrillator (ICD). Consistent biventricular pacing was limited by intermittent T-wave oversensing (TWOS). A strategy of left-ventricular-only pacing was used to eliminate TWOS. This strategy obviates the need to reduce ventricular sensitivity and thus may be an effective alternative to biventricular pacing complicated by TWOS. Copyright © 2013 Canadian Cardiovascular Society. Published by Elsevier Inc. All rights reserved.

  13. "Pseudo" Faraday cage: a solution for telemetry link interaction between a left ventricular assist device and an implantable cardioverter defibrillator.

    Science.gov (United States)

    Jacob, Sony; Cherian, Prasad K; Ghumman, Waqas S; Das, Mithilesh K

    2010-09-01

    Patients implanted with left ventricular assist devices (LVAD) may have implantable cardioverter defibrillators (ICD) implanted for sudden cardiac death prevention. This opens the possibility of device-device communication interactions and thus interferences. We present a case of such interaction that led to ICD communication failure following the activation of an LVAD. In this paper, we describe a practical solution to circumvent the communication interference and review the communication links of ICDs and possible mechanisms of ICD-LVAD interactions.

  14. External costs of electricity

    International Nuclear Information System (INIS)

    Rabl, A.; Spadaro, J.V.

    2005-01-01

    This article presents a synthesis of the ExternE project (External costs of Energy) of the European community about the external costs of power generation. Pollution impacts are calculated using an 'impact pathways' analysis, i.e. an analysis of the emission - dispersion - dose-response function - cost evaluation chain. Results are presented for different fuel cycles (with several technological variants) with their confidence intervals. The environmental impact costs are particularly high for coal: for instance, in France, for coal-fired power plants it is of the same order as the electricity retail price. For natural gas, this cost is about a third of the one for coal. On the contrary, the environmental impact costs for nuclear and renewable energies are low, typically of few per cent of the electricity price. The main part of these costs corresponds to the sanitary impacts, in particular the untimely mortality. In order to avoid any controversy about the cost evaluation of mortality, the reduction of the expectation of life due to the different fuel cycles is also indicated and the risks linked with nuclear energy are presented using several comparisons. (J.S.)

  15. On parabolic external maps

    DEFF Research Database (Denmark)

    Lomonaco, Luna; Petersen, Carsten Lunde; Shen, Weixiao

    2017-01-01

    We prove that any C1+BV degree d ≥ 2 circle covering h having all periodic orbits weakly expanding, is conjugate by a C1+BV diffeomorphism to a metrically expanding map. We use this to connect the space of parabolic external maps (coming from the theory of parabolic-like maps) to metrically expan...

  16. Stochastic Control - External Models

    DEFF Research Database (Denmark)

    Poulsen, Niels Kjølstad

    2005-01-01

    This note is devoted to control of stochastic systems described in discrete time. We are concerned with external descriptions or transfer function model, where we have a dynamic model for the input output relation only (i.e.. no direct internal information). The methods are based on LTI systems...

  17. A patient with severely reduced LV function and electrical storm saved by wearable cardioverter-defibrillator: a case report.

    Science.gov (United States)

    Strauss, Margit; Kouraki, Kleopatra; Skarlos, Alexandros; Zahn, Ralf; Kleemann, Thomas

    2013-06-01

    The wearable cardioverter-defibrillator (WCD) is indicated in patients who are considered to be at temporarily high risk for sudden cardiac death (SCD), when an implantable defibrillator is not yet clearly indicated. We report the case of a 41-year-old patient with a newly diagnosed severely reduced left ventricular (LV) function for suspected myocarditis and repeated nonsustained ventricular tachycardia (VT). This patient was supplied with a WCD who came back to the hospital 4 weeks after discharge with an electrical storm and adequate discharge of the WCD. After application of amiodarone, no further arrhythmias were detected during intrahospital course. For further risk stratification, we performed a magnetic field imaging (MFI), that was reported to be useful in risk assessment of SCD in patients with ischemic cardiomyopathy. This measurement showed a normal result, but we decided to give an implantable cardioverter-defibrillator (ICD) to the patient. During a follow-up of 1 year, no further arrhythmias occurred. With this case, we report the efficacy of a WCD, which is a novel tool in patients at temporarily high risk of SCD and we report a novel method of risk stratification in patients with a high risk of SCD.

  18. [Ethics in intensive care and euthanasia : With respect to inactivating defibrillators at the end of life in terminally ill patients].

    Science.gov (United States)

    Trappe, H-J

    2017-04-01

    In critically ill patients, intensive care medical procedures allow diseases to be cured or controlled that were considered incurable many years ago. For patients with terminal heart failure or heart disease with other severe comorbidities (cancer, stroke), the questions whether the deactivation of defibrillators is appropriate or must be regarded as active euthanasia may arise. Notable cases from the author's hospital are analyzed. The literature on the topic euthanasia and basic literature regarding defibrillator therapy are discussed. It is undisputed that patients as part of their self-determination have the right to renounce treatment. Active euthanasia and the thereby deliberate induction of death is prohibited by law in Germany and will be prosecuted. Passive euthanasia is the omission or reduction of possibly life-prolonging treatment measures. Passive euthanasia requires the patient's consent and is legally and ethically permissible. Indirect euthanasia takes into account acceleration of death as a side effect of a medication. Unpunishable assisted suicide ("assisted suicide") is the mere assistance of self-controlled and self-determined death. Assisted suicide is fundamentally not a criminal offense in Germany. Deactivation of a defibrillator is a treatment discontinuation, which is only permitted in accordance with the wishes of the patient. It is not a question of passive or active euthanasia. Involvement of a local ethics committee and/or legal consultation is certainly useful and sometimes also allows previously unrecognized questions to be answered.

  19. Using a Combined Platform of Swarm Intelligence Algorithms and GIS to Provide Land Suitability Maps for Locating Cardiac Rehabilitation Defibrillators

    Science.gov (United States)

    KAFFASH-CHARANDABI, Neda; SADEGHI-NIARAKI, Abolghasem; PARK, Dong-Kyun

    2015-01-01

    Background: Cardiac arrest is a condition in which the heart is completely stopped and is not pumping any blood. Although most cardiac arrest cases are reported from homes or hospitals, about 20% occur in public areas. Therefore, these areas need to be investigated in terms of cardiac arrest incidence so that places of high incidence can be identified and cardiac rehabilitation defibrillators installed there. Methods: In order to investigate a study area in Petersburg, Pennsylvania State, and to determine appropriate places for installing defibrillators with 5-year period data, swarm intelligence algorithms were used. Moreover, the location of the defibrillators was determined based on the following five evaluation criteria: land use, altitude of the area, economic conditions, distance from hospitals and approximate areas of reported cases of cardiac arrest for public places that were created in geospatial information system (GIS). Results: The A-P HADEL algorithm results were more precise about 27.36%. The validation results indicated a wider coverage of real values and the verification results confirmed the faster and more exact optimization of the cost function in the PSO method. Conclusion: The study findings emphasize the necessity of applying optimal optimization methods along with GIS and precise selection of criteria in the selection of optimal locations for installing medical facilities because the selected algorithm and criteria dramatically affect the final responses. Meanwhile, providing land suitability maps for installing facilities across hot and risky spots has the potential to save many lives. PMID:26587471

  20. Decalogue of electric defibrillation Decálogo de la desfibrilación eléctrica

    Directory of Open Access Journals (Sweden)

    Elkín Ferdinand Cardona Duque

    2001-03-01

    Full Text Available Defibrillation is an emergency procedure and the only effective therapy for ventricular fibrillation. Electrical defibrillation delivers large amounts of current to the myocardium and thus depolarizes it, terminating ventricular fibrillation and other arrhythmias. A defibrillator is a device that administers a controlled electrical shock, allowing the operator to select a variable current at the precise moment, according to patient‘s condition. Understanding defibrillator‘s operation leads to more effective resuscitation rates and more therapeutic alternatives in patients with any cardiac electric disturbance. La desfibrilación es un procedimiento de emergencia y es la única terapia efectiva para el manejo de la fibrilación ventricular. La desfibrilación eléctrica libera corriente en gran cantidad al miocardio, despolarizándolo y terminando la fibrilación ventricular y otras arritmias. Un desfibrilador es un aparato que suministra un choque eléctrico en forma controlada, permitiendo al operador seleccionar una corriente variable en el momento oportuno, de acuerdo con la condición del paciente. El entendimiento del manejo del desfibrilador permite tasas de resucitación más efectivas y más alternativas terapéuticas en pacientes con trastornos del ritmo cardíaco.

  1. An Automated Processing Method for Agglomeration Areas

    Directory of Open Access Journals (Sweden)

    Chengming Li

    2018-05-01

    Full Text Available Agglomeration operations are a core component of the automated generalization of aggregated area groups. However, because geographical elements that possess agglomeration features are relatively scarce, the current literature has not given sufficient attention to agglomeration operations. Furthermore, most reports on the subject are limited to the general conceptual level. Consequently, current agglomeration methods are highly reliant on subjective determinations and cannot support intelligent computer processing. This paper proposes an automated processing method for agglomeration areas. Firstly, the proposed method automatically identifies agglomeration areas based on the width of the striped bridging area, distribution pattern index (DPI, shape similarity index (SSI, and overlap index (OI. Next, the progressive agglomeration operation is carried out, including the computation of the external boundary outlines and the extraction of agglomeration lines. The effectiveness and rationality of the proposed method has been validated by using actual census data of Chinese geographical conditions in the Jiangsu Province.

  2. Internal and external validation of an ESTRO delineation guideline

    DEFF Research Database (Denmark)

    Eldesoky, Ahmed R.; Yates, Esben Svitzer; Nyeng, Tine B

    2016-01-01

    Background and purpose To internally and externally validate an atlas based automated segmentation (ABAS) in loco-regional radiation therapy of breast cancer. Materials and methods Structures of 60 patients delineated according to the ESTRO consensus guideline were included in four categorized...... and axillary nodal levels and poor agreement for interpectoral, internal mammary nodal regions and LADCA. Correcting ABAS significantly improved all the results. External validation of ABAS showed comparable results. Conclusions ABAS is a clinically useful tool for segmenting structures in breast cancer loco...

  3. Ascending-ramp biphasic waveform has a lower defibrillation threshold and releases less troponin I than a truncated exponential biphasic waveform.

    Science.gov (United States)

    Huang, Jian; Walcott, Gregory P; Ruse, Richard B; Bohanan, Scott J; Killingsworth, Cheryl R; Ideker, Raymond E

    2012-09-11

    We tested the hypothesis that the shape of the shock waveform affects not only the defibrillation threshold but also the amount of cardiac damage. Defibrillation thresholds were determined for 11 waveforms-3 ascending-ramp waveforms, 3 descending-ramp waveforms, 3 rectilinear first-phase biphasic waveforms, a Gurvich waveform, and a truncated exponential biphasic waveform-in 6 pigs with electrodes in the right ventricular apex and superior vena cava. The ascending, descending, and rectilinear waveforms had 4-, 8-, and 16-millisecond first phases and a 3.5-millisecond rectilinear second phase that was half the voltage of the first phase. The exponential biphasic waveform had a 60% first-phase and a 50% second-phase tilt. In a second study, we attempted to defibrillate after 10 seconds of ventricular fibrillation with a single ≈30-J shock (6 pigs successfully defibrillated with 8-millisecond ascending, 8-millisecond rectilinear, and truncated exponential biphasic waveforms). Troponin I blood levels were determined before and 2 to 10 hours after the shock. The lowest-energy defibrillation threshold was for the 8-milliseconds ascending ramp (14.6±7.3 J [mean±SD]), which was significantly less than for the truncated exponential (19.6±6.3 J). Six hours after shock, troponin I was significantly less for the ascending-ramp waveform (0.80±0.54 ng/mL) than for the truncated exponential (1.92±0.47 ng/mL) or the rectilinear waveform (1.17±0.45 ng/mL). The ascending ramp has a significantly lower defibrillation threshold and at ≈30 J causes 58% less troponin I release than the truncated exponential biphasic shock. Therefore, the shock waveform affects both the defibrillation threshold and the amount of cardiac damage.

  4. Automated Vehicles Symposium 2015

    CERN Document Server

    Beiker, Sven

    2016-01-01

    This edited book comprises papers about the impacts, benefits and challenges of connected and automated cars. It is the third volume of the LNMOB series dealing with Road Vehicle Automation. The book comprises contributions from researchers, industry practitioners and policy makers, covering perspectives from the U.S., Europe and Japan. It is based on the Automated Vehicles Symposium 2015 which was jointly organized by the Association of Unmanned Vehicle Systems International (AUVSI) and the Transportation Research Board (TRB) in Ann Arbor, Michigan, in July 2015. The topical spectrum includes, but is not limited to, public sector activities, human factors, ethical and business aspects, energy and technological perspectives, vehicle systems and transportation infrastructure. This book is an indispensable source of information for academic researchers, industrial engineers and policy makers interested in the topic of road vehicle automation.

  5. Hydrometeorological Automated Data System

    Data.gov (United States)

    National Oceanic and Atmospheric Administration, Department of Commerce — The Office of Hydrologic Development of the National Weather Service operates HADS, the Hydrometeorological Automated Data System. This data set contains the last 48...

  6. Planning for Office Automation.

    Science.gov (United States)

    Mick, Colin K.

    1983-01-01

    Outlines a practical approach to planning for office automation termed the "Focused Process Approach" (the "what" phase, "how" phase, "doing" phase) which is a synthesis of the problem-solving and participatory planning approaches. Thirteen references are provided. (EJS)

  7. Fixed automated spray technology.

    Science.gov (United States)

    2011-04-19

    This research project evaluated the construction and performance of Boschungs Fixed Automated : Spray Technology (FAST) system. The FAST system automatically sprays de-icing material on : the bridge when icing conditions are about to occur. The FA...

  8. Automated Vehicles Symposium 2014

    CERN Document Server

    Beiker, Sven; Road Vehicle Automation 2

    2015-01-01

    This paper collection is the second volume of the LNMOB series on Road Vehicle Automation. The book contains a comprehensive review of current technical, socio-economic, and legal perspectives written by experts coming from public authorities, companies and universities in the U.S., Europe and Japan. It originates from the Automated Vehicle Symposium 2014, which was jointly organized by the Association for Unmanned Vehicle Systems International (AUVSI) and the Transportation Research Board (TRB) in Burlingame, CA, in July 2014. The contributions discuss the challenges arising from the integration of highly automated and self-driving vehicles into the transportation system, with a focus on human factors and different deployment scenarios. This book is an indispensable source of information for academic researchers, industrial engineers, and policy makers interested in the topic of road vehicle automation.

  9. Automation Interface Design Development

    Data.gov (United States)

    National Aeronautics and Space Administration — Our research makes its contributions at two levels. At one level, we addressed the problems of interaction between humans and computers/automation in a particular...

  10. I-94 Automation FAQs

    Data.gov (United States)

    Department of Homeland Security — In order to increase efficiency, reduce operating costs and streamline the admissions process, U.S. Customs and Border Protection has automated Form I-94 at air and...

  11. Automation synthesis modules review

    International Nuclear Information System (INIS)

    Boschi, S.; Lodi, F.; Malizia, C.; Cicoria, G.; Marengo, M.

    2013-01-01

    The introduction of 68 Ga labelled tracers has changed the diagnostic approach to neuroendocrine tumours and the availability of a reliable, long-lived 68 Ge/ 68 Ga generator has been at the bases of the development of 68 Ga radiopharmacy. The huge increase in clinical demand, the impact of regulatory issues and a careful radioprotection of the operators have boosted for extensive automation of the production process. The development of automated systems for 68 Ga radiochemistry, different engineering and software strategies and post-processing of the eluate were discussed along with impact of automation with regulations. - Highlights: ► Generators availability and robust chemistry boosted for the huge diffusion of 68Ga radiopharmaceuticals. ► Different technological approaches for 68Ga radiopharmaceuticals will be discussed. ► Generator eluate post processing and evolution to cassette based systems were the major issues in automation. ► Impact of regulations on the technological development will be also considered

  12. Disassembly automation automated systems with cognitive abilities

    CERN Document Server

    Vongbunyong, Supachai

    2015-01-01

    This book presents a number of aspects to be considered in the development of disassembly automation, including the mechanical system, vision system and intelligent planner. The implementation of cognitive robotics increases the flexibility and degree of autonomy of the disassembly system. Disassembly, as a step in the treatment of end-of-life products, can allow the recovery of embodied value left within disposed products, as well as the appropriate separation of potentially-hazardous components. In the end-of-life treatment industry, disassembly has largely been limited to manual labor, which is expensive in developed countries. Automation is one possible solution for economic feasibility. The target audience primarily comprises researchers and experts in the field, but the book may also be beneficial for graduate students.

  13. Highway Electrification And Automation

    OpenAIRE

    Shladover, Steven E.

    1992-01-01

    This report addresses how the California Department of Transportation and the California PATH Program have made efforts to evaluate the feasibility and applicability of highway electrification and automation technologies. In addition to describing how the work was conducted, the report also describes the findings on highway electrification and highway automation, with experimental results, design study results, and a region-wide application impacts study for Los Angeles.

  14. Automated lattice data generation

    Directory of Open Access Journals (Sweden)

    Ayyar Venkitesh

    2018-01-01

    Full Text Available The process of generating ensembles of gauge configurations (and measuring various observables over them can be tedious and error-prone when done “by hand”. In practice, most of this procedure can be automated with the use of a workflow manager. We discuss how this automation can be accomplished using Taxi, a minimal Python-based workflow manager built for generating lattice data. We present a case study demonstrating this technology.

  15. Automated lattice data generation

    Science.gov (United States)

    Ayyar, Venkitesh; Hackett, Daniel C.; Jay, William I.; Neil, Ethan T.

    2018-03-01

    The process of generating ensembles of gauge configurations (and measuring various observables over them) can be tedious and error-prone when done "by hand". In practice, most of this procedure can be automated with the use of a workflow manager. We discuss how this automation can be accomplished using Taxi, a minimal Python-based workflow manager built for generating lattice data. We present a case study demonstrating this technology.

  16. Automated security management

    CERN Document Server

    Al-Shaer, Ehab; Xie, Geoffrey

    2013-01-01

    In this contributed volume, leading international researchers explore configuration modeling and checking, vulnerability and risk assessment, configuration analysis, and diagnostics and discovery. The authors equip readers to understand automated security management systems and techniques that increase overall network assurability and usability. These constantly changing networks defend against cyber attacks by integrating hundreds of security devices such as firewalls, IPSec gateways, IDS/IPS, authentication servers, authorization/RBAC servers, and crypto systems. Automated Security Managemen

  17. Marketing automation supporting sales

    OpenAIRE

    Sandell, Niko

    2016-01-01

    The past couple of decades has been a time of major changes in marketing. Digitalization has become a permanent part of marketing and at the same time enabled efficient collection of data. Personalization and customization of content are playing a crucial role in marketing when new customers are acquired. This has also created a need for automation to facilitate the distribution of targeted content. As a result of successful marketing automation more information of the customers is gathered ...

  18. Instant Sikuli test automation

    CERN Document Server

    Lau, Ben

    2013-01-01

    Get to grips with a new technology, understand what it is and what it can do for you, and then get to work with the most important features and tasks. A concise guide written in an easy-to follow style using the Starter guide approach.This book is aimed at automation and testing professionals who want to use Sikuli to automate GUI. Some Python programming experience is assumed.

  19. Managing laboratory automation

    OpenAIRE

    Saboe, Thomas J.

    1995-01-01

    This paper discusses the process of managing automated systems through their life cycles within the quality-control (QC) laboratory environment. The focus is on the process of directing and managing the evolving automation of a laboratory; system examples are given. The author shows how both task and data systems have evolved, and how they interrelate. A BIG picture, or continuum view, is presented and some of the reasons for success or failure of the various examples cited are explored. Fina...

  20. Shielded cells transfer automation

    International Nuclear Information System (INIS)

    Fisher, J.J.

    1984-01-01

    Nuclear waste from shielded cells is removed, packaged, and transferred manually in many nuclear facilities. Radiation exposure is absorbed by operators during these operations and limited only through procedural controls. Technological advances in automation using robotics have allowed a production waste removal operation to be automated to reduce radiation exposure. The robotic system bags waste containers out of glove box and transfers them to a shielded container. Operators control the system outside the system work area via television cameras. 9 figures

  1. Automated Status Notification System

    Science.gov (United States)

    2005-01-01

    NASA Lewis Research Center's Automated Status Notification System (ASNS) was born out of need. To prevent "hacker attacks," Lewis' telephone system needed to monitor communications activities 24 hr a day, 7 days a week. With decreasing staff resources, this continuous monitoring had to be automated. By utilizing existing communications hardware, a UNIX workstation, and NAWK (a pattern scanning and processing language), we implemented a continuous monitoring system.

  2. Automated Groundwater Screening

    International Nuclear Information System (INIS)

    Taylor, Glenn A.; Collard, Leonard B.

    2005-01-01

    The Automated Intruder Analysis has been extended to include an Automated Ground Water Screening option. This option screens 825 radionuclides while rigorously applying the National Council on Radiation Protection (NCRP) methodology. An extension to that methodology is presented to give a more realistic screening factor for those radionuclides which have significant daughters. The extension has the promise of reducing the number of radionuclides which must be tracked by the customer. By combining the Automated Intruder Analysis with the Automated Groundwater Screening a consistent set of assumptions and databases is used. A method is proposed to eliminate trigger values by performing rigorous calculation of the screening factor thereby reducing the number of radionuclides sent to further analysis. Using the same problem definitions as in previous groundwater screenings, the automated groundwater screening found one additional nuclide, Ge-68, which failed the screening. It also found that 18 of the 57 radionuclides contained in NCRP Table 3.1 failed the screening. This report describes the automated groundwater screening computer application

  3. ExternE transport methodology for external cost evaluation of air pollution

    DEFF Research Database (Denmark)

    Jensen, S. S.; Berkowicz, R.; Brandt, J.

    The report describes how the human exposure estimates based on NERI's human exposure modelling system (AirGIS) can improve the Danish data used for exposure factors in the ExternE Transport methodology. Initially, a brief description of the ExternE Tranport methodology is given and it is summarised...

  4. Slowing of electrical activity in ventricular fibrillation is not associated with increased defibrillation energies in the isolated rabbit heart

    Directory of Open Access Journals (Sweden)

    Jane eCaldwell

    2011-04-01

    Full Text Available Prolonged out-of-hospital ventricular fibrillation (VF arrests are associated with reduced ECG dominant frequency (DF and diminished defibrillation success. Partial reversal of ischaemia increases ECG DF and improves defibrillation outcome. We have investigated the metabolic components of ischaemia responsible for the decline in ECG DF and defibrillation success.Isolated Langendorff-perfused rabbit hearts were loaded with the voltage-sensitive dye RH237. Using a photodiode array, epicardial membrane potentials were recorded at 252 sites (15x15mm on the anterior surface of the left & right ventricles. Simultaneously, a global ECG was recorded. VF was induced by burst pacing, and after 60s, perfusion was either reduced to 6ml/min or the perfusate composition changed to impose hypoxia (95%N2/5%CO2, pH 6.7 (80%O2/20%CO2, or hyperkalaemia (8mM. Using Fast Fourier Transform, power spectra were created from the optical signals and the global ECG. The optical power spectra were summated to give a global power spectrum (pseudoECG. At 600s the minimum defibrillation voltage (MDV was determined by step-up protocol.During VF, the ECG and pseudoECG DF were reduced by low-flow ischaemia (9.0±1.0Hz, p<0.01, n=5 and raised [K+]o (12.2±1.3 Hz, p<0.05, n=7 compared to control (19.2±1.5 Hz, n=20, but were unaffected by acidic pHo (16.7±1.1 Hz, n=11 and hypoxia (14.0±1.2 Hz, n=10. In contrast, the MDV was raised by acidic pH (156.1±26.4V, p<0.001 and hypoxia (154.1±22.1V, p<0.01 compared to control (65.6±2.3V, but comparable changes were not observed in low-flow ischaemia (61.0±0.5V or raised [K+]o (56±3V. In summary, different metabolites are responsible for the reduction in DF and the increase in defibrillation energy during ischaemic VF.

  5. Chest compressions before defibrillation for out-of-hospital cardiac arrest: A meta-analysis of randomized controlled clinical trials

    Directory of Open Access Journals (Sweden)

    Meier Pascal

    2010-09-01

    Full Text Available Abstract Background Current 2005 guidelines for advanced cardiac life support strongly recommend immediate defibrillation for out-of-hospital cardiac arrest. However, findings from experimental and clinical studies have indicated a potential advantage of pretreatment with chest compression-only cardiopulmonary resuscitation (CPR prior to defibrillation in improving outcomes. The aim of this meta-analysis is to evaluate the beneficial effect of chest compression-first versus defibrillation-first on survival in patients with out-of-hospital cardiac arrest. Methods Main outcome measures were survival to hospital discharge (primary endpoint, return of spontaneous circulation (ROSC, neurologic outcome and long-term survival. Randomized, controlled clinical trials that were published between January 1, 1950, and June 19, 2010, were identified by a computerized search using SCOPUS, MEDLINE, BIOS, EMBASE, the Cochrane Central Register of Controlled Trials, International Pharmaceutical Abstracts database, and Web of Science and supplemented by conference proceedings. Random effects models were used to calculate pooled odds ratios (ORs. A subgroup analysis was conducted to explore the effects of response interval greater than 5 min on outcomes. Results A total of four trials enrolling 1503 subjects were integrated into this analysis. No difference was found between chest compression-first versus defibrillation-first in the rate of return of spontaneous circulation (OR 1.01 [0.82-1.26]; P = 0.979, survival to hospital discharge (OR 1.10 [0.70-1.70]; P = 0.686 or favorable neurologic outcomes (OR 1.02 [0.31-3.38]; P = 0.979. For 1-year survival, however, the OR point estimates favored chest compression first (OR 1.38 [0.95-2.02]; P = 0.092 but the 95% CI crossed 1.0, suggesting insufficient estimate precision. Similarly, for cases with prolonged response times (> 5 min point estimates pointed toward superiority of chest compression first (OR 1.45 [0

  6. Checklists for external validity

    DEFF Research Database (Denmark)

    Dyrvig, Anne-Kirstine; Kidholm, Kristian; Gerke, Oke

    2014-01-01

    to an implementation setting. In this paper, currently available checklists on external validity are identified, assessed and used as a basis for proposing a new improved instrument. METHOD: A systematic literature review was carried out in Pubmed, Embase and Cinahl on English-language papers without time restrictions....... The retrieved checklist items were assessed for (i) the methodology used in primary literature, justifying inclusion of each item; and (ii) the number of times each item appeared in checklists. RESULTS: Fifteen papers were identified, presenting a total of 21 checklists for external validity, yielding a total...... of 38 checklist items. Empirical support was considered the most valid methodology for item inclusion. Assessment of methodological justification showed that none of the items were supported empirically. Other kinds of literature justified the inclusion of 22 of the items, and 17 items were included...

  7. Participation through Automation: Fully Automated Critical PeakPricing in Commercial Buildings

    Energy Technology Data Exchange (ETDEWEB)

    Piette, Mary Ann; Watson, David S.; Motegi, Naoya; Kiliccote,Sila; Linkugel, Eric

    2006-06-20

    California electric utilities have been exploring the use of dynamic critical peak prices (CPP) and other demand response programs to help reduce peaks in customer electric loads. CPP is a tariff design to promote demand response. Levels of automation in DR can be defined as follows: Manual Demand Response involves a potentially labor-intensive approach such as manually turning off or changing comfort set points at each equipment switch or controller. Semi-Automated Demand Response involves a pre-programmed demand response strategy initiated by a person via centralized control system. Fully Automated Demand Response does not involve human intervention, but is initiated at a home, building, or facility through receipt of an external communications signal. The receipt of the external signal initiates pre-programmed demand response strategies. They refer to this as Auto-DR. This paper describes the development, testing, and results from automated CPP (Auto-CPP) as part of a utility project in California. The paper presents the project description and test methodology. This is followed by a discussion of Auto-DR strategies used in the field test buildings. They present a sample Auto-CPP load shape case study, and a selection of the Auto-CPP response data from September 29, 2005. If all twelve sites reached their maximum saving simultaneously, a total of approximately 2 MW of DR is available from these twelve sites that represent about two million ft{sup 2}. The average DR was about half that value, at about 1 MW. These savings translate to about 0.5 to 1.0 W/ft{sup 2} of demand reduction. They are continuing field demonstrations and economic evaluations to pursue increasing penetrations of automated DR that has demonstrated ability to provide a valuable DR resource for California.

  8. Analysis of implantable defibrillator longevity under clinical circumstances: implications for device selection.

    Science.gov (United States)

    Knops, Paul; Theuns, Dominic A M J; Res, Jan C J; Jordaens, Luc

    2009-10-01

    Information about implantable cardioverter-defibrillator (ICD) longevity is mostly calculated from measurements under ideal laboratory conditions. However, little information about longevity under clinical circumstances is available. This survey gives an overview on ICD service times and generator replacements in a cohort of consecutive ICD patients. Indications for replacement were classified as a normal end-of-service (EOS), premature EOS, system malfunction, infection and device advisory, or recall actions. From the premature and normal EOS group, longevity from single-chamber (SC), dual-chamber (DC), and cardiac resynchronization therapy defibrillator (CRT-D), rate-responsive (RR) settings, high output (HO) stimulation, and indication for ICD therapy was compared. Differences between brands were compared as well. In a total of 854 patients, 203 ICD replacements (165 patients) were recorded. Premature and normal EOS replacements consisted of 32 SC, 98 DC and 24 CRT-D systems. Longevity was significantly longer in SC systems compared to DC and CRT-D systems (54 +/- 19 vs. 40 +/- 17 and 42 +/- 15 months; P = 0.008). Longevity between non-RR (n = 143) and RR (n = 11) settings was not significantly different (43 +/- 18 vs. 45 +/- 13 months) as it also was not for HO versus non-HO stimulation (43 +/- 19 vs. 46 +/- 17 months). Longevity of ICDs was not significantly different between primary and secondary prevention (42 +/- 19 vs. 44 +/- 18 months). The average longevity on account of a device-based EOS message was 43 +/- 18 months. Average longevity for Biotronik (BIO, n = 72) was 33 +/- 10 months, for ELA Medical (ELA, n = 12) 44 +/- 17 months, for Guidant (GDT, n = 36) 49 +/- 12 months, for Medtronic (MDT, n = 29) 62 +/- 22 months, and for St. Jude Medical (SJM, n = 5) 31 +/- 9 months (P generators had a longer service time compared to DC and CRT-D systems. No influence of indication for ICD therapy and HO stimulation on generator longevity was observed in this

  9. Comparison of clinical benefits and outcome in patients with programmable and nonprogrammable implantable cardioverter defibrillators.

    Science.gov (United States)

    Mehta, D; Saksena, S; Krol, R B; Makhija, V

    1992-09-01

    Technological advances in implantable cardioverter defibrillators (ICDs) have provided a variety of programmable parameters and antitachycardia therapies whose utility and impact on clinical outcome is presently unknown. ICDs have capabilities for cardioversion defibrillation alone (first generation ICDs), or in conjunction with demand ventricular pacing (second generation ICDs), or with demand pacing and antitachycardia pacing (third generation ICDs). We examined the pattern of antitachycardia therapy use and long-term survival in 110 patients with sustained ventricular tachycardia (VT) or ventricular fibrillation (VF). Group I included 62 patients with nonprogrammable first generation ICDs that delivered committed shock therapy after ventricular tachyarrhythmia detection based on electrogram rate and/or morphology was satisfied. Group II included 48 patients with multiprogrammable ICDs (including second and third generation ICDs) that had programmable tachyarrhythmia detection based on rate and tachycardia confirmation prior to delivery of electrical treatment with either programmable shocks and/or, as in the third generation ICDs, antitachycardia pacing. Incidence and patterns of antitachycardia therapy use and long-term survival were compared in the two groups. The incidence of appropriate shocks in patients who completed 1 year of follow-up was significantly greater in group I (30 of 43 patients = 70% vs 11 of 26 patients = 42%; P less than 0.05). In the total follow-up period, a significantly larger proportion of group I patients as compared to group II patients used the shock therapies (46 of 62 patients = 74% vs 25 of 48 patients = 52%; P less than 0.01), with the majority doing so within the first year of implantation (96% and 92%, respectively). Although the frequency of antitachycardia therapy activation was similar, the number of shocks delivered per patient was lower in group II, particularly in the initial 3 months of follow-up (P = 0.06). No clinical

  10. AUTOMATION OF IMAGE DATA PROCESSING

    Directory of Open Access Journals (Sweden)

    Preuss Ryszard

    2014-12-01

    Full Text Available This article discusses the current capabilities of automate processing of the image data on the example of using PhotoScan software by Agisoft . At present, image data obtained by various registration systems (metric and non - metric cameras placed on airplanes , satellites , or more often on UAVs is used to create photogrammetric products. Multiple registrations of object or land area (large groups of photos are captured are usually performed in order to eliminate obscured area as well as to raise the final accuracy of the photogrammetric product. Because of such a situation t he geometry of the resulting image blocks is far from the typical configuration of images . For fast images georeferencing automatic image matching algorithms are currently applied . They can create a model of a block in the local coordinate system or using initial exterior orientation and measured control points can provide image georeference in an external reference frame. In the case of non - metric image application, it is also possible to carry out self - calibration process at this stage . Image matching algorithm is also used in generation of dense point clouds reconstructing spatial shape of the object ( area. In subsequent processing steps it is possible to obtain typical photogrammetric products such as orthomosaic , DSM or DTM and a photorealistic solid model of an object . All aforementioned processing steps are implemented in a single program in contrary to standard commercial software dividing all steps into dedicated modules . I mage processing leading to final geo referenced products can be fully automated including sequential implementation of the processing steps at predetermined control parameters . The paper presents the practical results of the application fully automatic generation of othomosaic for both images obtained by a metric Vexell camera and a block of images acquired by a non - metric UAV system.

  11. Pacemaker, implanted cardiac defibrillator and irradiation: Management proposal in 2010 depending on the type of cardiac stimulator and prognosis and location of cancer

    International Nuclear Information System (INIS)

    Lambert, P.; Da Costa, A.; Marcy, P.Y.; Kreps, S.; Angellier, G.; Marcie, S.; Bondiau, P.Y.; Briand-Amoros, C.; Thariat, J.

    2011-01-01

    Ionizing radiation may interfere with electric components of pacemakers or implantable cardioverter defibrillators. The type, severity and extent of radiation damage to pacemakers, have previously been shown to depend on the total dose and dose rate. Over 300,000 new cancer cases are treated yearly in France, among which 60% are irradiated in the course of their disease. One among 400 of these patients has an implanted pacemaker or defibrillator. The incidence of pacemaker and implanted cardioverter defibrillator increases in an ageing population. The oncologic prognosis must be weighted against the cardiologic prognosis in a multidisciplinary and transversal setting. Innovative irradiation techniques and technological sophistications of pacemakers and implantable cardioverter-defibrillators (with the introduction of more radiosensitive complementary metal-oxide-semiconductors since 1970) have potentially changed the tolerance profiles. This review of the literature studied the geometric, dosimetric and radiobiological characteristics of the radiation beams for high energy photons, stereotactic irradiation, proton-therapy. Standardized protocols and radiotherapy optimization (particle, treatment fields, energy) are advisable in order to improve patient management during radiotherapy and prolonged monitoring is necessary following radiation therapy. The dose received at the pacemaker/heart should be calculated. The threshold for the cumulated dose to the pacemaker/implantable cardioverter-defibrillator (2 to 5 Gy depending on the brand), the necessity to remove/displace the device based on the dose-volume histogram on dosimetry, as well as the use of lead shielding and magnet are discussed. (authors)

  12. Lithium-manganese dioxide cells for implantable defibrillator devices - Discharge voltage models

    Energy Technology Data Exchange (ETDEWEB)

    Root, Michael J. [Cardiac Rhythm Management Research and Development, Boston Scientific Corp., 4100 Hamline Ave. N., St. Paul, MN 55112 (United States)

    2010-08-01

    The discharge potential behavior of lithium-manganese dioxide cells designed for implantable cardiac defibrillators was characterized as a function of extent of cell depletion for tests designed to discharge the cells for times between 1 and 7 years. The discharge potential curves may be separated into two segments from 0 {<=} x {<=} {proportional_to}0.51 and {proportional_to}0.51 {<=} x {<=} 1.00, where x is the dimensionless extent of discharge referenced to the rated cell capacity. The discharge potentials conform to Tafel kinetics in each segment. This behavior allows the discharge potential curves to be predicted for an arbitrary discharge load and long term discharge performance may be predicted from short term test results. The discharge potentials may subsequently be modeled by fitting the discharge curves to empirical functions like polynomials and Pade approximants. A function based on the Nernst equation that includes a term accounting for nonideal interactions between lithium ions and the cathode host material, such as the Redlich-Kister relationship, also may be used to predict discharge behavior. (author)

  13. [Guidelines on the management of implantable cardioverter defibrillators at the end of life].

    Science.gov (United States)

    Datino, T; Rexach, L; Vidán, M T; Alonso, A; Gándara, Á; Ruiz-García, J; Fontecha, B; Martínez-Sellés, M

    2014-01-01

    This article is a joint document of the Spanish Society of Geriatrics and Gerontology, the Spanish Society of Palliative Care and the Section of Geriatric Cardiology of the Spanish Society of Cardiology. Its aim is to address the huge gap that exists in Spain with regard to the management of implantable cardioverter defibrillators (ICDs) in the final stages of life. It is increasingly common to find patients carrying these devices that are in the terminal stage of an advanced disease. This occurs in patients with advanced heart disease and subsequent heart failure refractory to treatment but also in a patient with an ICD who develops cancer disease, organ failure or other neurodegenerative diseases with poor short-term prognosis. The vast majority of these patients are over 65, so the paper focuses particularly on the elderly who are in this situation, but the decision-making process is similar in younger patients with ICDs who are in the final phase of their life. Copyright © 2013 SEGG. Published by Elsevier Espana. All rights reserved.

  14. Guidelines on the management of implantable cardioverter defibrillators at the end of life.

    Science.gov (United States)

    Datino, T; Rexach, L; Vidán, M T; Alonso, A; Gándara, Á; Ruiz-García, J; Fontecha, B; Martínez-Sellés, M

    2014-01-01

    This article is a joint document of the Spanish Society of Geriatrics and Gerontology, the Spanish Society of Palliative Care and the Section of Geriatric Cardiology of the Spanish Society of Cardiology. Its aim is to address the huge gap that exists in Spain with regard to the management of implantable cardioverter defibrillators (ICDs) in the final stages of life. It is increasingly common to find patients carrying these devices that are in the terminal stage of an advanced disease. This occurs in patients with advanced heart disease and subsequent heart failure refractory to treatment but also in a patient with an ICD who develops cancer disease, organ failure or other neurodegenerative diseases with poor short-term prognosis. The vast majority of these patients are over 65, so the paper focuses particularly on the elderly who are in this situation, but the decision-making process is similar in younger patients with ICDs who are in the final phase of their life. Copyright © 2013 Elsevier España, S.L. All rights reserved.

  15. Cardiac pacing systems and implantable cardiac defibrillators (ICDs): a radiological perspective of equipment, anatomy and complications

    Energy Technology Data Exchange (ETDEWEB)

    Burney, K. E-mail: apqz59@dsl.pipex.comk1511@hotmail.com; Burchard, F.; Papouchado, M.; Wilde, P

    2004-08-01

    Cardiac pacing is a proven and effective treatment in the management of many cardiac arrhythmias. Implantable cardiac defibrillators (ICDs) are beneficial for certain patient groups with a history of serious, recurrent ventricular dysrhythmias, with a high risk of sudden cardiac death. Pacemaker devices take many forms and are highly visible on the chest radiograph. The radiographic appearances of ICDs and pacemakers can be similar and are subject to similar complications. The anatomical approach to the implantation, the type of device used and anatomical variations will all affect the appearance of these devices on the chest film. Pacemaker complications identified radiographically include pneumothorax, lead malpositioning, lead displacement or fracture, fracture of outer conductor coil, loose connection between the lead and pacemaker connector block, lack of redundant loops in paediatric patients and excessive manipulation of the device by the patient (Twiddler's syndrome). This pictorial review highlights the role of chest radiography in the diagnosis of post-cardiac pacing and ICD insertion complications, as well as demonstrating the normal appearances of the most frequently implanted devices.

  16. Time dependence of risks and benefits in pediatric primary prevention implantable cardioverter-defibrillator therapy.

    Science.gov (United States)

    DeWitt, Elizabeth S; Triedman, John K; Cecchin, Frank; Mah, Doug Y; Abrams, Dominic J; Walsh, Edward P; Gauvreau, Kimberlee; Alexander, Mark E

    2014-12-01

    Implantable cardioverter defibrillators (ICDs) used to prevent sudden cardiac arrest in children not only provide appropriate therapy in 25% of patients but also result in a significant incidence of inappropriate shocks and other device complications. ICDs placed for secondary prevention have higher rates of appropriate therapy than those placed for primary prevention. Pediatric patients with primary prevention ICDs were studied to determine time-dependent incidence of appropriate use and adverse events. A total of 140 patients aged prevention were retrospectively identified. Demographics and times to first appropriate shock; adverse events (including inappropriate shock, lead failure, reintervention, and complication); generator replacement and follow-up were noted. During mean follow-up of 4 years, appropriate shock occurred in 19% patients and first adverse event (excluding death/transplant) occurred in 36%. Risk of death or transplant was ≈1% per year and was not related to receiving appropriate therapy. Conditional survival analysis showed rates of appropriate therapy and adverse events decrease soon after implantation, but adverse events are more frequent than appropriate therapy throughout follow-up. Primary prevention ICDs were associated with appropriate therapy in 19% and adverse event in 36% in this cohort. The incidence of both first appropriate therapy and device-related adverse events decreased during longer periods of follow-up after implantation. This suggests that indications for continued device therapy in pediatric primary prevention ICD patients might be reconsidered after a period of nonuse. © 2014 American Heart Association, Inc.

  17. [The daily experience of the patient with an implantable cardioverter defibrillator].

    Science.gov (United States)

    Palacios-Ceña, Domingo; Alonso-Blanco, Cristina; Cachón-Pérez, José Miguel; Alvarez-López, Cristina

    2010-01-01

    To describe the daily experience of patients with an automatic defibrillator (AD) implant and the adaptive changes of the patient. Qualitative and phenomenological research. Collection of data through; initially unstructured interview with half of the informants, semi-structured interviews through an open questions guide after the initial unstructured interviews and use of personal narratives of the informants. Analysis of the data using the Van Manen proposal. We analysed the interviews of 10 participants. We collected socio-demographic variables and identified the following themes, which respond to the question "How is life with an AD": It is life "with the two sides of the coin," living in constant wait and uncertainty, accepting change, developing adaptation strategies, renegotiating relationships and sexuality and it is to live transformed. The results of this study can be integrated into nurse clinical practice in areas such as valuation after discharge, changes in habits, control of treatment, notification of shocks, masking detection of symptoms and strategies that can jeopardise the bearer. Research needs to be developed that looks closer into the influence of other technological devices in people. Copyright 2009 Elsevier España, S.L. All rights reserved.

  18. Evidence of Cognitive Bias in Decision Making Around Implantable-Cardioverter Defibrillators: A Qualitative Framework Analysis.

    Science.gov (United States)

    Matlock, Daniel D; Jones, Jacqueline; Nowels, Carolyn T; Jenkins, Amy; Allen, Larry A; Kutner, Jean S

    2017-11-01

    Studies have demonstrated that patients with primary prevention implantable cardioverter-defibrillators (ICDs) often misunderstand the ICD. Advances in behavioral economics demonstrate that some misunderstandings may be due to cognitive biases. We aimed to explore the influence of cognitive bias on ICD decision making. We used a qualitative framework analysis including 9 cognitive biases: affect heuristic, affective forecasting, anchoring, availability, default effects, halo effects, optimism bias, framing effects, and state dependence. We interviewed 48 patients from 4 settings in Denver. The majority were male (n = 32). Overall median age was 61 years. We found frequent evidence for framing, default, and halo effects; some evidence of optimism bias, affect heuristic, state dependence, anchoring and availability bias; and little or no evidence of affective forecasting. Framing effects were apparent in overestimation of benefits and downplaying or omitting potential harms. We found evidence of cognitive bias in decision making for ICD implantation. The majority of these biases appeared to encourage ICD treatment. Published by Elsevier Inc.

  19. Congenital short QT syndrome and implantable cardioverter defibrillator treatment: inherent risk for inappropriate shock delivery.

    Science.gov (United States)

    Schimpf, Rainer; Wolpert, Christian; Bianchi, Francesca; Giustetto, Carla; Gaita, Florenzo; Bauersfeld, Urs; Borggrefe, Martin

    2003-12-01

    A congenital short QT interval constitutes a new primary electrical abnormality associated with syncope and/or sudden cardiac death. We report on the initial use of implantable cardioverter defibrillator (ICD) therapy in patients with inherited short QT interval and discuss sensing abnormalities and detection issues. In five consecutive patients from two unrelated European families who had structurally normal hearts, excessively shortened QT intervals, and a strong positive family history of sudden cardiac death, ICDs were placed for primary and secondary prevention. Mean QT intervals were 252 +/- 13 ms (QTc 287 +/- 13 ms). Despite normal sensing behavior during intraoperative and postoperative device testing, 3 of 5 patients experienced inappropriate shock therapies for T wave oversensing 30 +/- 26 days after implantation. Programming lower sensitivities and decay delays prevented further inappropriate discharges. The congenital short QT syndrome constitutes a new clinical entity with an increased risk for sudden cardiac death. Currently, ICD treatment is the only therapeutic option. In patients with short QT interval and implanted ICD, increased risk for inappropriate therapy is inherent due to the detection of short-coupled and prominent T waves. Careful testing of ICD function and adaptation of sensing levels and decay delays without sacrificing correct arrhythmia detection are essential.

  20. Interference of neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators: an in vitro study.

    Science.gov (United States)

    Ryf, Salome; Wolber, Thomas; Duru, Firat; Luechinger, Roger

    2008-01-01

    Permanent magnets may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets have become widely available in recent years and are incorporated in various articles of daily life. We conducted an in-vitro study to evaluate the ability of NdFeB magnets for home and office use to cause interference with cardiac pacemakers and ICDs. The magnetic fields of ten NdFeB magnets of different size and shape were measured at increasing distances beginning from the surface until a field-strength (B-field) value of 0.5 mT was reached. Furthermore, for each magnet the distance was determined at which a sample pacemaker switched from magnet mode to normal mode. Depending on the size and remanence of individual magnets, a B-field value of 0.5 mT was found at distances ranging from 1.5 cm to 30 cm and a value of 1 mT at distances from 1 cm to 22 cm. The pacemaker behavior was influenced at distances from 1 cm to 24 cm. NdFeB magnets for home and office use may cause interference with cardiac pacemakers and ICDs at distances up to 24 centimeters. Patient education and product declarations should include information about the risk associated with these magnets.

  1. Potential interference of small neodymium magnets with cardiac pacemakers and implantable cardioverter-defibrillators.

    Science.gov (United States)

    Wolber, Thomas; Ryf, Salome; Binggeli, Christian; Holzmeister, Johannes; Brunckhorst, Corinna; Luechinger, Roger; Duru, Firat

    2007-01-01

    Magnetic fields may interfere with the function of cardiac pacemakers and implantable cardioverter-defibrillators (ICDs). Neodymium-iron-boron (NdFeB) magnets, which are small in size but produce strong magnetic fields, have become widely available in recent years. Therefore, NdFeB magnets may be associated with an emerging risk of device interference. We conducted a clinical study to evaluate the potential of small NdFeB magnets to interfere with cardiac pacemakers and ICDs. The effect of four NdFeB magnets (two spherical magnets 8 and 10 mm in diameter, a necklace made of 45 spherical magnets, and a magnetic name tag) was tested in forty-one ambulatory patients with a pacemaker and 29 patients with an ICD. The maximum distance at which the magnetic switch of a device was influenced was observed. Magnetic interference was observed in all patients. The maximum distance resulting in device interference was 3 cm. No significant differences were found with respect to device manufacturer and device types. Small NdFeB magnets may cause interference with cardiac pacemakers and ICDs. Patients should be cautioned about the interference risk associated with NdFeB magnets during daily life.

  2. Electrical storm in patients with an implanted defibrillator: a matter of definition.

    Science.gov (United States)

    Israel, Carsten W; Barold, S Serge

    2007-10-01

    The term "electrical storm" (ES) indicates a state of cardiac electrical instability manifested by several episodes of ventricular tachyarrhythmias (VTs) within a short time. In patients with an implantable cardioverter defibrillator (ICD), ES is best defined as 3 appropriate VT detections in 24 h, treated by antitachycardia pacing, shock or eventually untreated but sustained in a VT monitoring zone. The number of shocks and inappropriate detections are irrelevant for the definition. ES occurs in approximately 25% of ICD patients within 3 years, with typically 5-55 individual VTs within one storm. Potential triggers can be found in approximately 66% of patients and include new/worsened heart failure, changes in antiarrhythmic medication, context with other illness, psychological stress, diarrhea, and hypokalemia. In most patients, ES consists of monomorphic VT indicating the presence of reentry while ventricular fibrillation indicating acute ischemia is rare. ES seems to have a low immediate mortality (1%) but frequently (50-80%) leads to hospitalization. Long-term prognostic implications of ES are unclear. The key intervention in ES is reduction of the elevated sympathetic tone by beta blockers and frequently benzodiazepines. Amiodarone i.v. has also been successful and azimilide seems promising while class I antiarrhythmic drugs are usually unsuccessful. Substrate mapping and VT ablation may be useful in treatment and prevention of ES. Prevention of ES requires ICD programming systematically avoiding unnecessary shocks (long VT detection, antitachycardia pacing where ever possible) which otherwise can fuel the sympathetic tone and prolong ES.

  3. Cardiac e-learning: Development of a web-based implantable cardioverter defibrillator educational system.

    Science.gov (United States)

    Hickey, Kathleen T; Johnson, Mary P; Biviano, Angelo; Aboelela, Sally; Thomas, Tami; Bakken, Suzanne; Garan, Hasan; Zimmerman, John L; Whang, William

    2011-04-01

    The objective of this study was to design a Web-based implantable cardioverter defibrillator (ICD) module that would allow greater access to learning which could occur at an individual's convenience outside the fast-paced clinical environment. A Web-based ICD software educational program was developed to provide general knowledge of the function of the ICD and the interpretation of the stored electrocardiograms. This learning tool could be accessed at any time via the Columbia University Internet server, using a unique, password protected login. A series of basic and advanced ICD terms were presented using actual ICD screenshots and videos that simulated scenarios the practitioner would most commonly encounter in the fast-paced clinical setting. To determine the usefulness of the site and improve the module, practitioners were asked to complete a brief (less than 5 min) online survey at the end of the module. Twenty-six practitioners have logged into our Web site: 20 nurses/nurse practitioners, four cardiac fellows, and two other practitioners. The majority of respondents rated the program as easy to use and useful. The success of this module has led to it becoming part of the training for student nurse practitioners before a clinical electrophysiology rotation, and the module is accessed by our cardiac entry level fellows before a rotation in the intensive care unit or electrophysiology service. Remote electronic arrhythmia learning is a successful example of the melding of technology and education to enhance clinical learning.

  4. Relation between total shock energy and mortality in patients with implantable cardioverter-defibrillator.

    Science.gov (United States)

    Tenma, Taro; Yokoshiki, Hisashi; Mitsuyama, Hirofumi; Watanabe, Masaya; Mizukami, Kazuya; Kamada, Rui; Takahashi, Masayuki; Sasaki, Ryo; Maeno, Motoki; Okamoto, Kaori; Chiba, Yuki; Anzai, Toshihisa

    2018-05-15

    Implantable Cardioverter-Defibrillator (ICD) shocks have been associated with mortality. However, no study has examined the relation between total shock energy and mortality. The aim of this study is to assess the association of total shock energy with mortality, and to determine the patients who are at risk of this association. Data from 316 consecutive patients who underwent initial ICD implantation in our hospital between 2000 and 2011 were retrospectively studied. We collected shock energy for 3 years from the ICD implantation, and determined the relation of shock energy on mortality after adjusting confounding factors. Eighty-seven ICD recipients experienced shock(s) within 3 years from ICD implantation and 43 patients had died during the follow-up. The amount of shock energy was significantly associated with all-cause death [adjusted hazard ratio (HR) 1.26 (per 100 joule increase), p energy accumulation (≥182 joule) was lower (p energy accumulation (energy accumulation and all-cause death was remarkable in the patients with low left ventricular ejection fraction (LVEF ≤40%) or atrial fibrillation (AF). Increase of shock energy was related to mortality in ICD recipients. This relation was evident in patients with low LVEF or AF. Copyright © 2018 Elsevier B.V. All rights reserved.

  5. Environmental external effects from wind power based on the EU ExternE methodology

    DEFF Research Database (Denmark)

    Ibsen, Liselotte Schleisner; Nielsen, Per Sieverts

    1998-01-01

    of the Danish part of the project is to implement the framework for externality evaluation, for three different power plants located in Denmark. The paper will focus on the assessment of the impacts of the whole fuel cycles for wind, natural gas and biogas. Priority areas for environmental impact assessment......The European Commission has launched a major study project, ExternE, to develop a methodology to quantify externalities. A “National Implementation Phase”, was started under the Joule II programme with the purpose of implementing the ExternE methodology in all member states. The main objective...

  6. Automation of Taxiing

    Directory of Open Access Journals (Sweden)

    Jaroslav Bursík

    2017-01-01

    Full Text Available The article focuses on the possibility of automation of taxiing, which is the part of a flight, which, under adverse weather conditions, greatly reduces the operational usability of an airport, and is the only part of a flight that has not been affected by automation, yet. Taxiing is currently handled manually by the pilot, who controls the airplane based on information from visual perception. The article primarily deals with possible ways of obtaining navigational information, and its automatic transfer to the controls. Analyzed wand assessed were currently available technologies such as computer vision, Light Detection and Ranging and Global Navigation Satellite System, which are useful for navigation and their general implementation into an airplane was designed. Obstacles to the implementation were identified, too. The result is a proposed combination of systems along with their installation into airplane’s systems so that it is possible to use the automated taxiing.

  7. External corners as heat bridges

    Energy Technology Data Exchange (ETDEWEB)

    Berber, J.

    1984-08-01

    The maximum additional heat loss in vertical external corners depending on wall thickness is determined. In order to amire at a low k-value, a much smaller wall thickness is required in externally insulated walls than in monolithic constructions; the greater loss of heat bridge with external insulation stands in contrast to a higher loss in thick, monolithic walls. In relation to total losses, the additional losses through external corners are practically negligible.

  8. Psychometric properties of HeartQoL, a core heart disease-specific health-related quality of life questionnaire, in Danish implantable cardioverter defibrillator recipients

    DEFF Research Database (Denmark)

    Zangger, Graziella; Zwisler, Ann-Dorthe; Kikkenborg Berg, Selina

    2018-01-01

    disease-specific health-related quality of life questionnaire, in implantable cardioverter defibrillator recipients. Design This study involved cross-sectional and test-retest study designs. Method Implantable cardioverter defibrillator recipients in the cross-sectional study completed the Heart......QoL, the Short-Form 36 Health Survey, and the Hospital Anxiety and Depression Scale. The HeartQoL structure, construct-related validity (convergent and discriminative) and reliability (internal consistency) were assessed. HeartQoL reproducibility (test-retest) was assessed in an independent sample of implantable...... psychometric attributes of validity and reliability in this implantable cardioverter defibrillator population. This study adds support for the HeartQoL as a core heart-specific health-related quality of life questionnaire in a broad group of patients with heart disease including implantable cardioverter...

  9. [External pancreatic fistulas management].

    Science.gov (United States)

    Stepan, E V; Ermolov, A S; Rogal', M L; Teterin, Yu S

    The main principles of treatment of external postoperative pancreatic fistulas are viewed in the article. Pancreatic trauma was the reason of pancreatic fistula in 38.7% of the cases, operations because of acute pancreatitis - in 25.8%, and pancreatic pseudocyst drainage - in 35.5%. 93 patients recovered after the treatment. Complex conservative treatment of EPF allowed to close fistulas in 74.2% of the patients with normal patency of the main pancreatic duct (MPD). The usage of octreotide 600-900 mcg daily for at least 5 days to decrease pancreatic secretion was an important part of the conservative treatment. Endoscopic papillotomy was performed in patients with major duodenal papilla obstruction and interruption of transporting of pancreatic secretion to duodenum. Stent of the main pancreatic duct was indicated in patients with extended pancreatic duct stenosis to normalize transport of pancreatic secretion to duodenum. Surgical formation of anastomosis between distal part of the main pancreatic duct and gastro-intestinal tract was carried out when it was impossible to fulfill endoscopic stenting of pancreatic duct either because of its interruption and diastasis between its ends, or in the cases of unsuccessful conservative treatment of external pancreatic fistula caused by drainage of pseudocyst.

  10. Control and automation systems

    International Nuclear Information System (INIS)

    Schmidt, R.; Zillich, H.

    1986-01-01

    A survey is given of the development of control and automation systems for energy uses. General remarks about control and automation schemes are followed by a description of modern process control systems along with process control processes as such. After discussing the particular process control requirements of nuclear power plants the paper deals with the reliability and availability of process control systems and refers to computerized simulation processes. The subsequent paragraphs are dedicated to descriptions of the operating floor, ergonomic conditions, existing systems, flue gas desulfurization systems, the electromagnetic influences on digital circuits as well as of light wave uses. (HAG) [de

  11. Automated nuclear materials accounting

    International Nuclear Information System (INIS)

    Pacak, P.; Moravec, J.

    1982-01-01

    An automated state system of accounting for nuclear materials data was established in Czechoslovakia in 1979. A file was compiled of 12 programs in the PL/1 language. The file is divided into four groups according to logical associations, namely programs for data input and checking, programs for handling the basic data file, programs for report outputs in the form of worksheets and magnetic tape records, and programs for book inventory listing, document inventory handling and materials balance listing. A similar automated system of nuclear fuel inventory for a light water reactor was introduced for internal purposes in the Institute of Nuclear Research (UJV). (H.S.)

  12. Automating the CMS DAQ

    International Nuclear Information System (INIS)

    Bauer, G; Darlea, G-L; Gomez-Ceballos, G; Bawej, T; Chaze, O; Coarasa, J A; Deldicque, C; Dobson, M; Dupont, A; Gigi, D; Glege, F; Gomez-Reino, R; Hartl, C; Hegeman, J; Masetti, L; Behrens, U; Branson, J; Cittolin, S; Holzner, A; Erhan, S

    2014-01-01

    We present the automation mechanisms that have been added to the Data Acquisition and Run Control systems of the Compact Muon Solenoid (CMS) experiment during Run 1 of the LHC, ranging from the automation of routine tasks to automatic error recovery and context-sensitive guidance to the operator. These mechanisms helped CMS to maintain a data taking efficiency above 90% and to even improve it to 95% towards the end of Run 1, despite an increase in the occurrence of single-event upsets in sub-detector electronics at high LHC luminosity.

  13. Altering user' acceptance of automation through prior automation exposure.

    Science.gov (United States)

    Bekier, Marek; Molesworth, Brett R C

    2017-06-01

    Air navigation service providers worldwide see increased use of automation as one solution to overcome the capacity constraints imbedded in the present air traffic management (ATM) system. However, increased use of automation within any system is dependent on user acceptance. The present research sought to determine if the point at which an individual is no longer willing to accept or cooperate with automation can be manipulated. Forty participants underwent training on a computer-based air traffic control programme, followed by two ATM exercises (order counterbalanced), one with and one without the aid of automation. Results revealed after exposure to a task with automation assistance, user acceptance of high(er) levels of automation ('tipping point') decreased; suggesting it is indeed possible to alter automation acceptance. Practitioner Summary: This paper investigates whether the point at which a user of automation rejects automation (i.e. 'tipping point') is constant or can be manipulated. The results revealed after exposure to a task with automation assistance, user acceptance of high(er) levels of automation decreased; suggesting it is possible to alter automation acceptance.

  14. Atrial electrogram quality in single-pass defibrillator leads with floating atrial bipole in patients with permanent atrial fibrillation and cardiac resynchronization therapy.

    Science.gov (United States)

    Sticherling, Christian; Müller, Dirk; Schaer, Beat A; Krüger, Silke; Kolb, Christof

    2018-03-27

    Many patients receiving cardiac resynchronization therapy (CRT) suffer from permanent atrial fibrillation (AF). Knowledge of the atrial rhythm is important to direct pharmacological or interventional treatment as well as maintaining AV-synchronous biventricular pacing if sinus rhythm can be restored. A single pass single-coil defibrillator lead with a floating atrial bipole has been shown to obtain reliable information about the atrial rhythm but has never been employed in a CRT-system. The purpose of this study was to assess the feasibility of implanting a single coil right ventricular ICD lead with a floating atrial bipole and the signal quality of atrial electrograms (AEGM) in CRT-defibrillator recipients with permanent AF. Seventeen patients (16 males, mean age 73 ± 6 years, mean EF 25 ± 5%) with permanent AF and an indication for CRT-defibrillator placement were implanted with a designated CRT-D system comprising a single pass defibrillator lead with a atrial floating bipole. They were followed-up for 103 ± 22 days using remote monitoring for AEGM transmission. All patients had at last one AEGM suitable for atrial rhythm diagnosis and of 100 AEGM 99% were suitable for visual atrial rhythm assessment. Four patients were discharged in sinus rhythm and one reverted to AF during follow-up. Atrial electrograms retrieved from a single-pass defibrillator lead with a floating atrial bipole can be reliably used for atrial rhythm diagnosis in CRT recipients with permanent AF. Hence, a single pass ventricular defibrillator lead with a floating bipole can be considered in this population. Copyright © 2018 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. All rights reserved.

  15. Prompt prediction of successful defibrillation from 1-s ventricular fibrillation waveform in patients with out-of-hospital sudden cardiac arrest.

    Science.gov (United States)

    Endoh, Hiroshi; Hida, Seiji; Oohashi, Satomi; Hayashi, Yusuke; Kinoshita, Hidenori; Honda, Tadayuki

    2011-02-01

    Ventricular fibrillation (VF) is a common cardiac arrest rhythm that can be terminated by electrical defibrillation. During cardiopulmonary resuscitation, there is a strong need for a prompt and reliable predictor of successful defibrillation because myocardial damage can result from repeated futile defibrillation attempts. Continuous wavelet transform (CWT) provides excellent time and frequency resolution of signals. The purpose of this study was to evaluate whether features based on CWT could predict successful defibrillation. VF electrocardiogram (ECG) waveforms stored in ambulance-located defibrillators were collected. Predefibrillation waveforms were divided into 1.0- or 5.12-s VF waveforms. Indices in frequency domain or nonlinear analysis were calculated on the 5.12-s waveform. Simultaneously, CWT was performed on the 1.0-s waveform, and total low-band (1-3 Hz), mid-band (3-10 Hz), and high-band (10-32 Hz) energy were calculated. In 152 patients with out-of-hospital cardiac arrest, a total of 233 ECG predefibrillation recordings, consisting of 164 unsuccessful and 69 successful episodes, were analyzed. Indices of frequency domain analysis (peak frequency, centroid frequency, and amplitude spectral area), nonlinear analysis (approximate entropy and Hurst exponent, detrended fluctuation analysis), and CWT analysis (mid-band and high-band energy) were significantly different between unsuccessful and successful episodes (P centroid frequency and total mid-band energy were effective predictors (P < 0.01 for both). Energy spectrum analysis based on CWT as short as a 1.0-s VF ECG waveform enables prompt and reliable prediction of successful defibrillation.

  16. LIBRARY AUTOMATION IN NIGERAN UNIVERSITIES

    African Journals Online (AJOL)

    facilitate services and access to information in libraries is widely acceptable. ... Moreover, Ugah (2001) reports that the automation process at the. Abubakar ... blueprint in 1987 and a turn-key system of automation was suggested for the library.

  17. Cost-effectiveness of a risk-stratified approach to cardiac resynchronisation therapy defibrillators (high versus low) at the time of generator change.

    Science.gov (United States)

    Claridge, Simon; Sebag, Frederic A; Fearn, Steven; Behar, Jonathan M; Porter, Bradley; Jackson, Tom; Sieniewicz, Benjamin; Gould, Justin; Webb, Jessica; Chen, Zhong; O'Neill, Mark; Gill, Jaswinder; Leclercq, Christophe; Rinaldi, Christopher A

    2018-03-01

    Responders to cardiac resynchronisation therapy whose device has a defibrillator component and who do not receive a therapy in the lifetime of the first generator have a very low incidence of appropriate therapy after box change. We investigated the cost implications of using a risk stratification tool at the time of generator change resulting in these patients being reimplanted with a resynchronisation pacemaker. A decision tree was created using previously published data which had demonstrated an annualised appropriate defibrillator therapy risk of 2.33%. Costs were calculated at National Health Service (NHS) national tariff rates (2016-2017). EQ-5D utility values were applied to device reimplantations, admissions and mortality data, which were then used to estimate quality-adjusted life-years (QALYs) over 5 years. At 5 years, the incremental cost of replacing a resynchronisation defibrillator device with a second resynchronisation defibrillator versus resynchronisation pacemaker was £5045 per patient. Incremental QALY gained was 0.0165 (defibrillator vs pacemaker), resulting in an incremental cost-effectiveness ratio (ICER) of £305 712 per QALYs gained. Probabilistic sensitivity analysis resulted in an ICER of £313 612 (defibrillator vs pacemaker). For reimplantation of all patients with a defibrillator rather than a pacemaker to yield an ICER of less than £30 000 per QALY gained (current NHS cut-off for approval of treatment), the annual arrhythmic event rate would need to be 9.3%. The budget impact of selective replacement was a saving of £2 133 985 per year. Implanting low-risk patients with a resynchronisation defibrillator with the same device at the time of generator change is not cost-effective by current NHS criteria. Further research is required to understand the impact of these findings on individual patients at the time of generator change. © Article author(s) (or their employer(s) unless otherwise stated in the text of the article) 2018

  18. Future Trends in Process Automation

    OpenAIRE

    Jämsä-Jounela, Sirkka-Liisa

    2007-01-01

    The importance of automation in the process industries has increased dramatically in recent years. In the highly industrialized countries, process automation serves to enhance product quality, master the whole range of products, improve process safety and plant availability, efficiently utilize resources and lower emissions. In the rapidly developing countries, mass production is the main motivation for applying process automation. The greatest demand for process automation is in the chemical...

  19. Adaptive Automation Design and Implementation

    Science.gov (United States)

    2015-09-17

    with an automated system to a real-world adaptive au- tomation system implementation. There have been plenty of adaptive automation 17 Adaptive...of systems without increasing manpower requirements by allocating routine tasks to automated aids, improving safety through the use of au- tomated ...between intermediate levels of au- tomation , explicitly defining which human task a given level automates. Each model aids the creation and classification

  20. Current use of implantable electrical devices in Sweden: data from the Swedish pacemaker and implantable cardioverter-defibrillator registry.

    Science.gov (United States)

    Gadler, Fredrik; Valzania, Cinzia; Linde, Cecilia

    2015-01-01

    The National Swedish Pacemaker and Implantable Cardioverter-Defibrillator (ICD) Registry collects prospective data on all pacemaker and ICD implants in Sweden. We aimed to report the 2012 findings of the Registry concerning electrical devices implantation rates and changes over time, 1 year complications, long-term device longevity and patient survival. Forty-four Swedish implanting centres continuously contribute implantation of pacemakers and ICDs to the Registry by direct data entry on a specific website. Clinical and technical information on 2012 first implants and postoperative complications were analysed and compared with previous years. Patient survival data were obtained from the Swedish population register database. In 2012, the mean pacemaker and ICD first implantation rates were 697 and 136 per million inhabitants, respectively. The number of cardiac resynchronization therapy (CRT) first implantations/million capita was 41 (CRT pacemakers) and 55 (CRT defibrillators), with only a slight increase in CRT-ICD rate compared with 2011. Most device implantations were performed in men. Complication rates for pacemaker and ICD procedures were 5.3 and 10.1% at 1 year, respectively. Device and lead longevity differed among manufacturers. Pacemaker patients were older at the time of first implant and had generally worse survival rate than ICD patients (63 vs. 82% after 5 years). Pacemaker and ICD implantation rates seem to have reached a level phase in Sweden. Implantable cardioverter-defibrillator and CRT implantation rates are very low and do not reflect guideline indications. Gender differences in CRT and ICD implantations are pronounced. Device and patient survival rates are variable, and should be considered when deciding device type. Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2014. For permissions please email: journals.permissions@oup.com.

  1. External Measures of Cognition

    Directory of Open Access Journals (Sweden)

    Osvaldo eCairo

    2011-10-01

    Full Text Available The human brain is undoubtedly the most impressive, complex and intricate organ that has evolved over time. It is also probably the least understood, and for that reason, the one that is currently attracting the most attention. In fact, the number of comparative analyses that focus on the evolution of brain size in Homo sapiens and other species has increased dramatically in recent years. In neuroscience, no other issue has generated so much interest and been the topic of so many heated debates as the difference in brain size between socially defined population groups, both its connotations and implications. For over a century, external measures of cognition have been related to intelligence. However, it is still unclear whether these measures actually correspond to cognitive abilities. In summary, this paper must be reviewed with this premise in mind.

  2. Hanford External Dosimetry Program

    International Nuclear Information System (INIS)

    Fix, J.J.

    1990-10-01

    This document describes the Hanford External Dosimetry Program as it is administered by Pacific Northwest Laboratory (PNL) in support of the US Department of Energy (DOE) and its Hanford contractors. Program services include administrating the Hanford personnel dosimeter processing program and ensuring that the related dosimeter data accurately reflect occupational dose received by Hanford personnel or visitors. Specific chapters of this report deal with the following subjects: personnel dosimetry organizations at Hanford and the associated DOE and contractor exposure guidelines; types, characteristics, and procurement of personnel dosimeters used at Hanford; personnel dosimeter identification, acceptance testing, accountability, and exchange; dosimeter processing and data recording practices; standard sources, calibration factors, and calibration processes (including algorithms) used for calibrating Hanford personnel dosimeters; system operating parameters required for assurance of dosimeter processing quality control; special dose evaluation methods applied for individuals under abnormal circumstances (i.e., lost results, etc.); and methods for evaluating personnel doses from nuclear accidents. 1 ref., 14 figs., 5 tabs

  3. Bevalac external beamline optics

    Energy Technology Data Exchange (ETDEWEB)

    Kalnins, J.G.; Krebs, G.F.; Tekawa, M.M.; Alonso, J.R.

    1987-04-01

    This handbook is intended as an aid for tuning the external particle beam (EPB) lines at the Lawrence Berkeley Laboratory's Bevalac. The information contained within will be useful to the Bevalac's Main Control Room and experimenters alike. First, some general information is given concerning the EPB lines and beam optics. Next, each beam line is described in detail: schematics of the beam line components are shown, all the variables required to run a beam transport program are presented, beam envelopes are given with wire chamber pictures and magnet currents, focal points and magnifications. Some preliminary scaling factors are then presented which should aid in choosing a given EPB magnet's current for a given central Bevalac field. Finally, some tuning hints are suggested.

  4. Bevalac external beamline optics

    International Nuclear Information System (INIS)

    Kalnins, J.G.; Krebs, G.F.; Tekawa, M.M.; Alonso, J.R.

    1987-04-01

    This handbook is intended as an aid for tuning the external particle beam (EPB) lines at the Lawrence Berkeley Laboratory's Bevalac. The information contained within will be useful to the Bevalac's Main Control Room and experimenters alike. First, some general information is given concerning the EPB lines and beam optics. Next, each beam line is described in detail: schematics of the beam line components are shown, all the variables required to run a beam transport program are presented, beam envelopes are given with wire chamber pictures and magnet currents, focal points and magnifications. Some preliminary scaling factors are then presented which should aid in choosing a given EPB magnet's current for a given central Bevalac field. Finally, some tuning hints are suggested

  5. Automated HAZOP revisited

    DEFF Research Database (Denmark)

    Taylor, J. R.

    2017-01-01

    Hazard and operability analysis (HAZOP) has developed from a tentative approach to hazard identification for process plants in the early 1970s to an almost universally accepted approach today, and a central technique of safety engineering. Techniques for automated HAZOP analysis were developed...

  6. Automated Student Model Improvement

    Science.gov (United States)

    Koedinger, Kenneth R.; McLaughlin, Elizabeth A.; Stamper, John C.

    2012-01-01

    Student modeling plays a critical role in developing and improving instruction and instructional technologies. We present a technique for automated improvement of student models that leverages the DataShop repository, crowd sourcing, and a version of the Learning Factors Analysis algorithm. We demonstrate this method on eleven educational…

  7. Automated Vehicle Monitoring System

    OpenAIRE

    Wibowo, Agustinus Deddy Arief; Heriansyah, Rudi

    2014-01-01

    An automated vehicle monitoring system is proposed in this paper. The surveillance system is based on image processing techniques such as background subtraction, colour balancing, chain code based shape detection, and blob. The proposed system will detect any human's head as appeared at the side mirrors. The detected head will be tracked and recorded for further action.

  8. Mechatronic Design Automation

    DEFF Research Database (Denmark)

    Fan, Zhun

    successfully design analogue filters, vibration absorbers, micro-electro-mechanical systems, and vehicle suspension systems, all in an automatic or semi-automatic way. It also investigates the very important issue of co-designing plant-structures and dynamic controllers in automated design of Mechatronic...

  9. Automated Accounting. Instructor Guide.

    Science.gov (United States)

    Moses, Duane R.

    This curriculum guide was developed to assist business instructors using Dac Easy Accounting College Edition Version 2.0 software in their accounting programs. The module consists of four units containing assignment sheets and job sheets designed to enable students to master competencies identified in the area of automated accounting. The first…

  10. Automated conflict resolution issues

    Science.gov (United States)

    Wike, Jeffrey S.

    1991-01-01

    A discussion is presented of how conflicts for Space Network resources should be resolved in the ATDRSS era. The following topics are presented: a description of how resource conflicts are currently resolved; a description of issues associated with automated conflict resolution; present conflict resolution strategies; and topics for further discussion.

  11. Automated gamma counters

    International Nuclear Information System (INIS)

    Regener, M.

    1977-01-01

    This is a report on the most recent developments in the full automation of gamma counting in RIA, in particular by Messrs. Kontron. The development targets were flexibility in sample capacity and shape of test tubes, the possibility of using different radioisotopes for labelling due to an optimisation of the detector system and the use of microprocessers to substitute software for hardware. (ORU) [de

  12. Myths in test automation

    Directory of Open Access Journals (Sweden)

    Jazmine Francis

    2014-12-01

    Full Text Available Myths in automation of software testing is an issue of discussion that echoes about the areas of service in validation of software industry. Probably, the first though that appears in knowledgeable reader would be Why this old topic again? What's New to discuss the matter? But, for the first time everyone agrees that undoubtedly automation testing today is not today what it used to be ten or fifteen years ago, because it has evolved in scope and magnitude. What began as a simple linear scripts for web applications today has a complex architecture and a hybrid framework to facilitate the implementation of testing applications developed with various platforms and technologies. Undoubtedly automation has advanced, but so did the myths associated with it. The change in perspective and knowledge of people on automation has altered the terrain. This article reflects the points of views and experience of the author in what has to do with the transformation of the original myths in new versions, and how they are derived; also provides his thoughts on the new generation of myths.

  13. Myths in test automation

    Directory of Open Access Journals (Sweden)

    Jazmine Francis

    2015-01-01

    Full Text Available Myths in automation of software testing is an issue of discussion that echoes about the areas of service in validation of software industry. Probably, the first though that appears in knowledgeable reader would be Why this old topic again? What's New to discuss the matter? But, for the first time everyone agrees that undoubtedly automation testing today is not today what it used to be ten or fifteen years ago, because it has evolved in scope and magnitude. What began as a simple linear scripts for web applications today has a complex architecture and a hybrid framework to facilitate the implementation of testing applications developed with various platforms and technologies. Undoubtedly automation has advanced, but so did the myths associated with it. The change in perspective and knowledge of people on automation has altered the terrain. This article reflects the points of views and experience of the author in what has to do with the transformation of the original myths in new versions, and how they are derived; also provides his thoughts on the new generation of myths.

  14. Building Automation Systems.

    Science.gov (United States)

    Honeywell, Inc., Minneapolis, Minn.

    A number of different automation systems for use in monitoring and controlling building equipment are described in this brochure. The system functions include--(1) collection of information, (2) processing and display of data at a central panel, and (3) taking corrective action by sounding alarms, making adjustments, or automatically starting and…

  15. Automation of activation analysis

    International Nuclear Information System (INIS)

    Ivanov, I.N.; Ivanets, V.N.; Filippov, V.V.

    1985-01-01

    The basic data on the methods and equipment of activation analysis are presented. Recommendations on the selection of activation analysis techniques, and especially the technique envisaging the use of short-lived isotopes, are given. The equipment possibilities to increase dataway carrying capacity, using modern computers for the automation of the analysis and data processing procedure, are shown

  16. Protokoller til Home Automation

    DEFF Research Database (Denmark)

    Kjær, Kristian Ellebæk

    2008-01-01

    computer, der kan skifte mellem foruddefinerede indstillinger. Nogle gange kan computeren fjernstyres over internettet, så man kan se hjemmets status fra en computer eller måske endda fra en mobiltelefon. Mens nævnte anvendelser er klassiske indenfor home automation, er yderligere funktionalitet dukket op...

  17. Automation of radioimmunoassays

    International Nuclear Information System (INIS)

    Goldie, D.J.; West, P.M.; Ismail, A.A.A.

    1979-01-01

    A short account is given of recent developments in automation of the RIA technique. Difficulties encountered in the incubation, separation and quantitation steps are summarized. Published references are given to a number of systems, both discrete and continuous flow, and details are given of a system developed by the present authors. (U.K.)

  18. Microcontroller for automation application

    Science.gov (United States)

    Cooper, H. W.

    1975-01-01

    The description of a microcontroller currently being developed for automation application was given. It is basically an 8-bit microcomputer with a 40K byte random access memory/read only memory, and can control a maximum of 12 devices through standard 15-line interface ports.

  19. Driver Psychology during Automated Platooning

    NARCIS (Netherlands)

    Heikoop, D.D.

    2017-01-01

    With the rapid increase in vehicle automation technology, the call for understanding how humans behave while driving in an automated vehicle becomes more urgent. Vehicles that have automated systems such as Lane Keeping Assist (LKA) or Adaptive Cruise Control (ACC) not only support drivers in their

  20. Risk of pacemaker or implantable cardioverter defibrillator after radiotherapy for early-stage breast cancer in Denmark, 1982-2005

    DEFF Research Database (Denmark)

    Rehammar, Jens Christian; Johansen, Jens Brock; Jensen, Maj-Britt

    2017-01-01

    BACKGROUND AND PURPOSE: To examine the risk of cardiac conduction abnormalities or severe ventricular arrhythmias requiring implantation of a cardiac implantable electronic device (CIED), either a pacemaker or an implantable cardioverter-defibrillator, subsequent to breast cancer (BC) radiotherapy...... (RT). MATERIAL AND METHODS: All women treated for early-stage BC in Denmark from 1982 to 2005 were identified from the Danish Breast Cancer Cooperative Group. By record linkage to the Danish Pacemaker and ICD Registry information was retrieved on CIED implants subsequent to RT. Standardized incidence...

  1. Cognitive behavioral therapy (CBT in a Patient with Implantable Cardioverter Defibrillator (ICD and Posttraumatic stress disorder(PTSD.

    Directory of Open Access Journals (Sweden)

    Sahar Ansari

    2014-09-01

    Full Text Available The implantable cardioverter defibrillator (ICD has currently become the standard treatment for preventing sudden cardiac death. There are some psychological consequences in patients with ICD such as posttraumatic stress disorder (PTSD after the shocks induced by ICD. This report aimed to present the case of a 54-year-old man with ICD who had developed PTSD; his PTSD was treated, using cognitive-behavioral psychotherapy consisting of relaxation, mindfulness and problem solving techniques. In patients with ICD who are experiencing PTSD using cognitive behavioral interventions may be helpful to reduce their psychological sufferings.

  2. Automated Inadvertent Intruder Application

    International Nuclear Information System (INIS)

    Koffman, Larry D.; Lee, Patricia L.; Cook, James R.; Wilhite, Elmer L.

    2008-01-01

    The Environmental Analysis and Performance Modeling group of Savannah River National Laboratory (SRNL) conducts performance assessments of the Savannah River Site (SRS) low-level waste facilities to meet the requirements of DOE Order 435.1. These performance assessments, which result in limits on the amounts of radiological substances that can be placed in the waste disposal facilities, consider numerous potential exposure pathways that could occur in the future. One set of exposure scenarios, known as inadvertent intruder analysis, considers the impact on hypothetical individuals who are assumed to inadvertently intrude onto the waste disposal site. Inadvertent intruder analysis considers three distinct scenarios for exposure referred to as the agriculture scenario, the resident scenario, and the post-drilling scenario. Each of these scenarios has specific exposure pathways that contribute to the overall dose for the scenario. For the inadvertent intruder analysis, the calculation of dose for the exposure pathways is a relatively straightforward algebraic calculation that utilizes dose conversion factors. Prior to 2004, these calculations were performed using an Excel spreadsheet. However, design checks of the spreadsheet calculations revealed that errors could be introduced inadvertently when copying spreadsheet formulas cell by cell and finding these errors was tedious and time consuming. This weakness led to the specification of functional requirements to create a software application that would automate the calculations for inadvertent intruder analysis using a controlled source of input parameters. This software application, named the Automated Inadvertent Intruder Application, has undergone rigorous testing of the internal calculations and meets software QA requirements. The Automated Inadvertent Intruder Application was intended to replace the previous spreadsheet analyses with an automated application that was verified to produce the same calculations and

  3. Automating spectral measurements

    Science.gov (United States)

    Goldstein, Fred T.

    2008-09-01

    This paper discusses the architecture of software utilized in spectroscopic measurements. As optical coatings become more sophisticated, there is mounting need to automate data acquisition (DAQ) from spectrophotometers. Such need is exacerbated when 100% inspection is required, ancillary devices are utilized, cost reduction is crucial, or security is vital. While instrument manufacturers normally provide point-and-click DAQ software, an application programming interface (API) may be missing. In such cases automation is impossible or expensive. An API is typically provided in libraries (*.dll, *.ocx) which may be embedded in user-developed applications. Users can thereby implement DAQ automation in several Windows languages. Another possibility, developed by FTG as an alternative to instrument manufacturers' software, is the ActiveX application (*.exe). ActiveX, a component of many Windows applications, provides means for programming and interoperability. This architecture permits a point-and-click program to act as automation client and server. Excel, for example, can control and be controlled by DAQ applications. Most importantly, ActiveX permits ancillary devices such as barcode readers and XY-stages to be easily and economically integrated into scanning procedures. Since an ActiveX application has its own user-interface, it can be independently tested. The ActiveX application then runs (visibly or invisibly) under DAQ software control. Automation capabilities are accessed via a built-in spectro-BASIC language with industry-standard (VBA-compatible) syntax. Supplementing ActiveX, spectro-BASIC also includes auxiliary serial port commands for interfacing programmable logic controllers (PLC). A typical application is automatic filter handling.

  4. Management of patients with implantable cardioverter-defibrillators and pacemakers who require radiation therapy.

    Science.gov (United States)

    Brambatti, Michela; Mathew, Rebecca; Strang, Barbara; Dean, Joan; Goyal, Anuja; Hayward, Joseph E; Long, Laurene; DeMeis, Patty; Smoke, Marcia; Connolly, Stuart J; Morillo, Carlos A; Amit, Guy; Capucci, Alessandro; Healey, Jeff S

    2015-10-01

    Radiation therapy (RT) may pose acute and long-term risks for patients with cardiac implantable electronic devices (CIEDs), including pacemakers (PMs) and implantable cardioverter-defibrillators (ICDs). However, the frequency of these problems has not been accurately defined. The purpose of this study was to determine the prevalence of CIEDs among patients requiring RT and report the common CIED-related problems when patients are managed according to a standard clinical care path. In a single tertiary-care center, we prospectively screened all patients requiring RT and identified patients with ICDs or PMs. We collected clinical data about their cancer, RT treatment plan, and CIED. Radiation dose to the device was estimated in all patients, and any device malfunction during RT was documented. Of the 34,706 consecutive patients receiving RT, 261 patients (0.8%, mean age 77.9 ± 9.4 years) had an implantable cardiac device: 54 (20.7%) ICDs and 207 (79.3%) PMs. The site of RT was head and neck (27.4%), chest (30.0%), and abdomen/pelvis (32.6%). Using our care path, 63.2% of patients required continuous cardiac monitoring, 14.6% required device reprogramming, 18.8% required magnet application during RT, and 3.4% required device repositioning to the contralateral side before RT. Four patients (1.5%) had inappropriate device function during RT: 3 experienced hemodynamically tolerated ventricular pacing at the maximum sensor rate, and 1 experienced a device power-on-reset. No patient died or suffered permanent device failure. Nearly 1% of patients receiving RT in this series has a PM or ICD. However, with a systematic policy of risk assessment and patient management, significant device-related complications are rare. Copyright © 2015 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

  5. Safety and interaction of patients with implantable cardiac defibrillators driving a hybrid vehicle.

    Science.gov (United States)

    Tondato, Fernando; Bazzell, Jane; Schwartz, Linda; Mc Donald, Bruce W; Fisher, Robert; Anderson, S Shawn; Galindo, Arcenio; Dueck, Amylou C; Scott, Luis R

    2017-01-15

    Electromagnetic interference (EMI) can affect the function of implantable cardioverter defibrillators (ICD). Hybrid electric vehicles (HEV) have increased popularity and are a potential source of EMI. Little is known about the in vivo effects of EMI generated by HEV on ICD. This study evaluated the in vivo interaction between EMI generated by HEV with ICD. Thirty patients (73±9 y/o; 80% male) with stable ICD function were exposed to EMI generated by a Toyota Prius Hybrid®. The vehicle was lifted above the ground, allowing safe changes in engine rotation and consequent variations in electromagnetic emission. EMI was measured (NARDA STS® model EHP-50C) and expressed in A/m (magnetic), Volts/m (electrical), and Hertz (frequency). Six positions were evaluated: driver, front passenger, right and left back seats, outside, at the back and front of the car. Each position was evaluated at idle, 30 mph, 60 mph and variable speeds (acceleration-deceleration-brake). All ICD devices were continuously monitored during the study. The levels of EMI generated were low (highest mean levels: 2.09A/m at right back seat at 30 mph; and 3.5V/m at driver seat at variable speeds). No episode of oversensing or inadvertent change in ICD programming was observed. It is safe for patients with ICD to interact with HEV. This is the first study to address this issue using an in vivo model. Further studies are necessary to evaluate the interaction of different models of HEV or electric engine with ICD or unipolar pacemakers. Copyright © 2016 Elsevier Ireland Ltd. All rights reserved.

  6. Influence of radiotherapy on the latest generation of implantable cardioverter-defibrillators

    International Nuclear Information System (INIS)

    Hurkmans, Coen W.; Scheepers, Egon; Springorum, Bob G.F.; Uiterwaal, Hans

    2005-01-01

    Purpose: Radiotherapy can influence the functioning of pacemakers and implantable cardioverter-defibrillators (ICDs). ICDs offer the same functionality as pacemakers, but are also able to deliver a high-voltage shock to the heart if needed. Guidelines for radiotherapy treatment of patients with an implanted rhythm device have been published in 1994 by The American Association of Physicists in Medicine, and are based only on experience with pacemakers. Data on the influence of radiotherapy on ICDs are limited. The objective of our study is to determine the influence of radiotherapy on the latest generation of ICDs. Methods and Materials: Eleven modern ICDs have been irradiated in our department. The irradiation was performed with a 6-MV photon beam. The given dose was fractionated up to a cumulative dose of 120 Gy. Two to 5 days passed between consecutive irradiations. Frequency, output, sensing, telemetry, and shock energy were monitored. Results: Sensing interference by ionizing radiation on all ICDs has been demonstrated. For four ICDs, this would have caused the inappropriate delivery of a shock because of interference. At the end of the irradiation sessions, all devices had reached their point of failure. Complete loss of function was observed for four ICDs at dose levels between 0.5 Gy and 1.5 Gy. Conclusions: The effect of radiation therapy on the newest generation of ICDs varies widely. If tachycardia monitoring and therapy are functional (programmed on) during irradiation, the ICD might inappropriately give antitachycardia therapy, often resulting in a shock. Although most ICDs did not fail below 80 Gy, some devices had already failed at doses below 1.5 Gy. Guidelines are formulated for the treatment of patients with an ICD

  7. Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator.

    Science.gov (United States)

    Russo, Robert J; Costa, Heather S; Silva, Patricia D; Anderson, Jeffrey L; Arshad, Aysha; Biederman, Robert W W; Boyle, Noel G; Frabizzio, Jennifer V; Birgersdotter-Green, Ulrika; Higgins, Steven L; Lampert, Rachel; Machado, Christian E; Martin, Edward T; Rivard, Andrew L; Rubenstein, Jason C; Schaerf, Raymond H M; Schwartz, Jennifer D; Shah, Dipan J; Tomassoni, Gery F; Tominaga, Gail T; Tonkin, Allison E; Uretsky, Seth; Wolff, Steven D

    2017-02-23

    The presence of a cardiovascular implantable electronic device has long been a contraindication for the performance of magnetic resonance imaging (MRI). We established a prospective registry to determine the risks associated with MRI at a magnetic field strength of 1.5 tesla for patients who had a pacemaker or implantable cardioverter-defibrillator (ICD) that was "non-MRI-conditional" (i.e., not approved by the Food and Drug Administration for MRI scanning). Patients in the registry were referred for clinically indicated nonthoracic MRI at a field strength of 1.5 tesla. Devices were interrogated before and after MRI with the use of a standardized protocol and were appropriately reprogrammed before the scanning. The primary end points were death, generator or lead failure, induced arrhythmia, loss of capture, or electrical reset during the scanning. The secondary end points were changes in device settings. MRI was performed in 1000 cases in which patients had a pacemaker and in 500 cases in which patients had an ICD. No deaths, lead failures, losses of capture, or ventricular arrhythmias occurred during MRI. One ICD generator could not be interrogated after MRI and required immediate replacement; the device had not been appropriately programmed per protocol before the MRI. We observed six cases of self-terminating atrial fibrillation or flutter and six cases of partial electrical reset. Changes in lead impedance, pacing threshold, battery voltage, and P-wave and R-wave amplitude exceeded prespecified thresholds in a small number of cases. Repeat MRI was not associated with an increase in adverse events. In this study, device or lead failure did not occur in any patient with a non-MRI-conditional pacemaker or ICD who underwent clinically indicated nonthoracic MRI at 1.5 tesla, was appropriately screened, and had the device reprogrammed in accordance with the prespecified protocol. (Funded by St. Jude Medical and others; MagnaSafe ClinicalTrials.gov number, NCT

  8. Diagnosis and therapy of atrial tachyarrhythmias in the dual chamber implantable cardioverter defibrillator.

    Science.gov (United States)

    Dijkman, B; Wellens, H J

    2000-11-01

    Devices capable of monitoring and treating atrial tachyarrhythmias provide information about the natural history of the arrhythmias and potentially can influence their natural course by electrical therapy early after onset. Types of atrial arrhythmias and efficacy of device therapies were evaluated in 30 patients implanted with the Medtronic model 7250 Jewel AF implantable cardioverter defibrillator (ICD). All patients had structural heart disease and documented sustained ventricular and atrial arrhythmias (27 with atrial fibrillation [AF]) before implant. Twenty patients were taking amiodarone, and three were taking sotalol. During 20+/-10 months of follow-up, 600 atrial arrhythmia recurrences were documented in 50% of patients. AF was diagnosed in 19%, fast polymorphic atrial tachycardia (AT) in 20%, fast monomorphic AT in 57%, and slow AT in 4% of episodes. The two adaptive pacing therapies, burst and ramp, together with the 50-Hz burst, were successful in 57% of detected atrial arrhythmias. Burst and ramp were responsible for 49% and 50-Hz burst for 51% of successfully treated arrhythmias; 33% of the episodes terminated spontaneously. No ventricular proarrhythmia was observed due to atrial pacing therapies. In 30% of episodes, dual chamber pacing was required due to post termination bradycardia. Atrial arrhythmia recurrences in patients with dilated cardiomyopathy were not amenable to pacing therapies. Several aspects of atrial arrhythmia diagnosis, therapy, and documentation that are specific for functioning of the Jewel AF are discussed. Atrial arrhythmias in ICD patients with diseased hearts who are taking Class III antiarrhythmics frequently had longer cycle lengths than AF. Half of these arrhythmias could be terminated with pacing therapies; one third terminated spontaneously.

  9. Importance of the atrial channel for ventricular arrhythmia therapy in the dual chamber implantable cardioverter defibrillator.

    Science.gov (United States)

    Dijkman, B; Wellens, H J

    2000-12-01

    Performance of dual chamber implantable cardioverter defibrillator (ICD) systems has been judged based on functioning of the ventricular tachycardia:supraventricular tachycardia (VT:SVT) discrimination criteria and DDD pacing. The purpose of this study was to evaluate the use of dual chamber diagnostics to improve the electrical and antiarrhythmic therapy of ventricular arrhythmias. Information about atrial and ventricular rhythm in relation to ventricular arrhythmia occurrence and therapy was evaluated in 724 spontaneous arrhythmia episodes detected and treated by three types of dual chamber ICDs in 41 patients with structural heart disease. Device programming was based on clinically documented and induced ventricular arrhythmias. In ambulatory patients, sinus tachycardia preceded ventricular arrhythmias more often than in the hospital during exercise testing. The incidence of these VTs could be reduced by increasing the dose of a beta-blocking agent in only two patients. In five patients in whom sinus tachycardia developed after onset of hemodynamic stable VT, propranolol was more effective than Class III antiarrhythmics combined with another beta-blocking agent with regard to the incidence of VT and pace termination. In all but three cases, atrial arrhythmias were present for a longer time before the onset of ventricular arrhythmias. During atrial arrhythmias, fast ventricular rates before the onset of ventricular rate were observed more often than RR irregularities and short-long RR sequences. Dual chamber diagnostics allowed proper interpretation of detection and therapy outcome in patients with different types of ventricular arrhythmia. The advantages of the dual chamber ICD system go further than avoiding the shortcomings of the single chamber system. Information from the atrial chamber allows better device programming and individualization of drug therapy for ventricular arrhythmia.

  10. Optimizing implantable cardioverter-defibrillator treatment of rapid ventricular tachycardia: antitachycardia pacing therapy during charging.

    Science.gov (United States)

    Schoels, Wolfgang; Steinhaus, David; Johnson, W Ben; O'hara, Gilles; Schwab, Joerg O; Jenniskens, Inge; Degroot, Paul J; Tang, Feng; Helmling, Erhard

    2007-07-01

    Previous studies in implantable cardioverter-defibrillator (ICD) patients demonstrated the efficacy and safety of antitachycardia pacing (ATP) for rapid ventricular tachycardias (VT). To prevent shock delay in case of ATP failure, a new feature (ATP during charging) was developed to deliver ATP for rapid VT while charging for shock. The purpose of this study was to determine the efficacy and safety of this new feature. In a prospective, nonrandomized trial, patients with standard ICD indication received an EnTrust ICD. VT and ventricular fibrillation (VF) episodes were reviewed for appropriate detection, ATP success, rhythm acceleration, and related symptoms. In 421 implanted patients, 116 VF episodes occurred in 37 patients. Eighty-four (72%) episodes received ATP during or before charging. ATP prevented a shock in 58 (69%) of 84 episodes in 15 patients. ATP stopped significantly more monomorphic (77%) than polymorphic VTs (44%, P = .05). Five (6%) episodes accelerated after ATP but were terminated by the backup shock(s). No symptoms were related to ATP during charging. In four patients, 38 charges were saved by delivering ATP before charging. Of 98 induced VF episodes, 28% were successfully terminated by ATP versus 69% for spontaneous episodes (P <.01). Most VTs detected in the VF zone can be painlessly terminated by ATP delivered during charging, with a low risk of acceleration or symptoms. ATP before charging allows delivery of two ATP attempts before shock in the same time that would otherwise be required to deliver only one ATP plus a shock. It also offers potential battery energy savings.

  11. Complications and Mortality of Single Versus Dual Chamber Implantable Cardioverter Defibrillators

    Directory of Open Access Journals (Sweden)

    Ataallah Bagherzadeh

    2006-04-01

    Full Text Available Background: The implantable cardioverter defibrillators (ICDs are increasingly being used as a treatment modality for life threatening tachyarrhythmia. The purpose of this study was to compare the frequency of complications and mortality between single-chamber and dual-chamber ICD implantation in Shahid Rajaie cardiovascular center. Methods and results: Between January 2000 and December 2004, 234 patients received ICD by a percutaneous transvenous approach and were followed for 33 ± 23 months. The cumulative incidence of complications was 9.4% over the follow-up period. There was no significant difference in overall complication rate between single chamber (VR and dual chamber (DR ICD groups in the follow-up period (P= 0.11. The risk of complications did not have any statistically significant difference in secondary versus primary prevention groups (P=0.06. The complications were not associated with the severity of left ventricular systolic dysfunction (P=0.16.The frequency of lead-related complications was higher in dual chamber ICDs in comparison with single chamber ICDs (P=0.02. There was no significant difference in mortality between different sex groups (P=0.37, different indications for ICD implantation (P=0.43 or between VR and DR ICD groups (P= 0.55. Predictors of mortality were NYHA class III or more (P65 years (P=0.011 and LVEF<30% (P<0.001. The mortality in patients with CAD and DCM were significantly higher than those with other structural heart diseases (P=0.001. Conclusions: Close monitoring of patients during the first 2 month after ICD implantation is recommended because the majority of complications occur early after the procedure.

  12. Digoxin Is Associated With Increased Shock Events and Electrical Storms in Patients With Implantable Cardioverter Defibrillators.

    Science.gov (United States)

    Mina, George S; Acharya, Madan; Shepherd, Taylor; Gobrial, George; Tekeste, Michael; Watti, Hussam; Bhandari, Ruchi; Saini, Aditya; Reddy, Pratap; Dominic, Paari

    2018-03-01

    Recently, digoxin use has been found to associate with higher mortality. Yet, potential mechanisms by which digoxin use increases mortality remain unclear. Increased arrhythmogenicity from digoxin use is one possibility. Thus, we aimed to evaluate the relation between digoxin and shock events in patients with implantable cardioverter defibrillators (ICDs). We performed a retrospective chart review of all patients with ICDs and at least 1 device interrogation at our institution between January 1, 2012, and January 1, 2015. We aimed to cover 1 year of interrogation period. Patients with heart failure, atrial fibrillation, or both were included in the analysis. Patients were divided into 2 groups based on digoxin use, defined as use of digoxin for any period of time during ICD interrogation period. Incidence of ICD shock events and electrical storms and hospitalizations were compared between the 2 groups. The study included 202 patients. Of those, 55 patients were on digoxin and 147 were not on digoxin. Patients on digoxin were more likely to receive ICD shocks (odds ratio [OR] = 2.5, 95% confidence interval [95% CI] = 1.01-6.18, P = .04) and have increased risk of electrical storms ( P = .02). Moreover, total hospitalizations were higher in digoxin users ( P = .02). Multivariate logistic regression analysis also showed that digoxin use was an independent predictor of shock events (OR = 4.07, 95% CI = 1.43-11.58, P = .009). Digoxin is associated with increased shock events and electrical storms in patients with ICDs; however, large randomized controlled studies are needed to confirm our findings.

  13. Long-term single-center experience of defibrillator therapy in children and adolescents.

    Science.gov (United States)

    Frommeyer, Gerrit; Feder, Sebastian; Bettin, Markus; Debus, Volker; Köbe, Julia; Reinke, Florian; Uebing, Anselm; Eckardt, Lars; Kehl, Hans Gerd

    2018-06-01

    Implantable cardioverter-defibrillator (ICD) systems are established therapy for prevention of sudden cardiac death. Long-term data on ICD systems in children and adolescents is rare. The present study displays a long-term single-center follow-up of children and adolescents with ICD systems. The present study represents a single-center experience of patients younger than 18 years who received an ICD (n = 58). Follow-up data included in-house follow-up as well as examinations of collaborating specialists. Mean age at implantation was 14.0 ± 3.3 years and 33 patients (56.9%) were male. A transvenous ICD system was implanted in 54 patients (93.1%). In 33 patients (56.9%) electrical heart disease or idiopathic ventricular fibrillation represented the underlying condition of ICD implantation. Median follow-up duration was 70 months (45; 94). 3 patients (5.2%) died during the observation period. None of these deaths was associated with ICD failure. Appropriate shocks occurred in 32 patients (55.2%). Inappropriate shock delivery was recorded in 17 patients (29.3%). Supraventricular tachycardia represented the most frequent cause of inappropriate shock delivery (9 patients, 52.9%). T-wave oversensing led to inappropriate shock delivery in 3 patients (17.6%). In 5 patients (29.4%), lead failure caused inappropriate shock delivery. Of note, during follow-up lead failure was reported in 15 patients (25.9%) leading to surgical revision. ICD therapy in children and adolescents is effective for prevention of sudden cardiac death. The rate of appropriate shock deliveries was significantly higher as compared with large ICD trials. Inappropriate therapies occurred frequently. In particular supraventricular tachycardia, T-wave oversensing and lead failures were responsible for these episodes. Copyright © 2017 Elsevier B.V. All rights reserved.

  14. Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failure.

    Science.gov (United States)

    Gunderson, Bruce D; Gillberg, Jeffrey M; Wood, Mark A; Vijayaraman, Pugazhendhi; Shepard, Richard K; Ellenbogen, Kenneth A

    2006-02-01

    Implantable cardioverter-defibrillator (ICD) lead failures often present as inappropriate shock therapy. An algorithm that can reliably discriminate between ventricular tachyarrhythmias and noise due to lead failure may prevent patient discomfort and anxiety and avoid device-induced proarrhythmia by preventing inappropriate ICD shocks. The goal of this analysis was to test an ICD tachycardia detection algorithm that differentiates noise due to lead failure from ventricular tachyarrhythmias. We tested an algorithm that uses a measure of the ventricular intracardiac electrogram baseline to discriminate the sinus rhythm isoelectric line from the right ventricular coil-can (i.e., far-field) electrogram during oversensing of noise caused by a lead failure. The baseline measure was defined as the product of the sum (mV) and standard deviation (mV) of the voltage samples for a 188-ms window centered on each sensed electrogram. If the minimum baseline measure of the last 12 beats was algorithm to detect lead failures. The minimum baseline measure for the 24 lead failure episodes (0.28 +/- 0.34 mV-mV) was smaller than the 135 ventricular tachycardia (40.8 +/- 43.0 mV-mV, P <.0001) and 55 ventricular fibrillation episodes (19.1 +/- 22.8 mV-mV, P <.05). A minimum baseline <0.35 mV-mV threshold had a sensitivity of 83% (20/24) with a 100% (190/190) specificity. A baseline measure of the far-field electrogram had a high sensitivity and specificity to detect lead failure noise compared with ventricular tachycardia or fibrillation.

  15. Adherence to an Aerobic Exercise Intervention after an Implantable Cardioverter Defibrillator (ICD).

    Science.gov (United States)

    Dougherty, Cynthia M; Luttrell, Matilda N; Burr, Robert L; Kim, Misun; Haskell, William L

    2016-02-01

    Exercise adherence is an important element in achieving important exercise outcomes. The purpose of this study was to describe adherence in a home-based aerobic exercise program following an implantable cardioverter defibrillator (ICD), determine effects of adherence on peakVO2 , and outline reasons for nonadherence. A single-blind randomized control trial of home walking compared to usual care in 160 patients with an ICD for primary or secondary prevention was conducted. This report is on adherence in the exercise arm (N = 84). Home walking exercise consisted of 8 weeks of aerobic conditioning (60 minutes/day, 5 days/week) followed by 16 weeks of aerobic maintenance (150 minutes/week, 30 minutes/session) at 60-80% of heart rate reserve. Adherence was tracked using Polar heart rate (HR) monitors, pedometers, home exercise logs, and telephone follow-up. Adherence was defined as performing at least 80% of prescribed exercise. For aerobic conditioning, there was a mean frequency of 3.81 walks/week, duration of 1,873 minutes walked, and 17.5% of exercise was in the target HR (THR) zone. For aerobic maintenance, there was a mean frequency of 2.4 walks/week, duration of 1,872 minutes/walked, and 8.7% of exercise was in the THR zone. Those who were 80% adherent achieved a 3.4 mL/kg/min (P = 0.03) improvement in peakVO2 over those who were exercise ranged from scheduling issues to viral illness and fatigue. Adherence to aerobic exercise frequency and duration was high with few dropouts, resulting in higher peakVO2 . Exercise monitoring equipment encouraged adherence and conferred a sense of safety to exercise. ©2015 Wiley Periodicals, Inc.

  16. Prognosis after implantation of cardioverter-defibrillators in Korean patients with Brugada syndrome.

    Science.gov (United States)

    Son, Myoung Kyun; Byeon, Kyeongmin; Park, Seung-Jung; Kim, June Soo; Nam, Gi-Byoung; Choi, Kee-Joon; Kim, You-Ho; Park, Sang Weon; Kim, Young-Hoon; Park, Hyung Wook; Cho, Jeong Gwan; On, Young Keun

    2014-01-01

    Our study aims to analyze prognosis after implantable cardioverter-defibrillator (ICD) implantation in Korean patients with Brugada syndrome (BrS). This was a retrospective study of BrS patients implanted with an ICD at one of four centers in Korea between January 1998 and April 2012. Sixty-nine patients (68 males, 1 female) were implanted with an ICD based on aborted cardiac arrest (n=38, 55%), history of syncope (n=17, 25%), or induced ventricular tachyarrhythmia on electrophysiologic study in asymptomatic patients (n=14, 20%). A family history of sudden cardiac death and a spontaneous type 1 electrocardiography (ECG) were noted in 13 patients (19%) and 44 patients (64%), respectively. During a mean follow-up of 59±46 months, 4.6±5.5 appropriate shocks were delivered in 19 patients (28%). Fourteen patients (20%) experienced 5.2±8.0 inappropriate shocks caused by supraventricular arrhythmia, lead failure, or abnormal sensing. Six patients were admitted for cardiac causes during follow-up, but no cardiac deaths occurred. An episode of aborted cardiac arrest was a significant predictor of appropriate shock, and the composite of cardiac events in the Cox proportional hazard model [hazard ratio (95% confidence interval) was 11.34 (1.31-97.94) and 4.78 (1.41-16.22), respectively]. However, a spontaneous type 1 ECG was not a predictor of cardiac events. Appropriate shock (28%) and inappropriate shock (20%) were noted during a mean follow-up of 59±46 months in Korean BrS patients implanted with an ICD. An episode of aborted cardiac arrest was the most powerful predictor of cardiac events.

  17. Infrequent physician use of implantable cardioverter-defibrillators risks patient safety.

    Science.gov (United States)

    Lyman, Stephen; Sedrakyan, Art; Do, Huong; Razzano, Renee; Mushlin, Alvin I

    2011-10-01

    Implantable cardioverter-defibrillators (ICDs) have diffused rapidly into clinical practice with little evaluation of their real-world effectiveness. To determine the effect of the adoption of ICD on patient safety, particularly with respect to physician volume and early outcomes. Retrospective cohort of all ICD implantations in New York state from 1997 to 2006, with follow-up at 90 days and 1 year. Setting New York state non-federal hospital discharges in which an ICD was implanted during the admission. Patients were followed forward for 1 year for subsequent admissions. Patients New York state residents undergoing ICD implantation. Effects of annual and career ICD implantation volume on 90-day complication, readmission, reprogramming, mortality and revision of the ICD within 1 year. This cohort (N = 38,992) represents a period of rapid adoption and implementation of this new technology, with frequency more than tripling between 1997 and 2006. We identified 6439 (16.5%) post-implantation complications and 1093 (2.8%) deaths within 90 days of implantation. The majority (73.4%) of physicians implanted one or fewer ICDs per year, and 11.0% of all implantations were performed by these very-low-volume operators. Patients treated by very-low-volume operators were more likely to die (RR = 1.8, 95% CI 1.3 to 2.4) or experience cardiac complications (RR = 4.7, 95% CI 3.3 to 6.8) even after the adjustment for case mix compared to operators who frequently performed ICD implantation. These findings suggest a need for safe and effective implementation strategies for new medical technologies, which minimize patient risk due to rapid diffusion among inexperienced providers and assure that the intended benefit can be maximised rapidly.

  18. Low risk of electromagnetic interference between smartphones and contemporary implantable cardioverter defibrillators.

    Science.gov (United States)

    Burri, Haran; Mondouagne Engkolo, Louis Paulin; Dayal, Nicolas; Etemadi, Abdul; Makhlouf, Anne-Marie; Stettler, Carine; Trentaz, Florence

    2016-05-01

    Manufacturers of implantable cardioverter defibrillators (ICDs) recommend that cell phones be maintained at a distance of ∼15 cm from the implanted device in order to avoid risk of dysfunction due to electromagnetic interference (EMI). Data relating to this issue are outdated and do not reflect modern technology. Our aim was to evaluate whether EMI is still an issue with contemporary ICDs and smartphones. Consecutive patients implanted with a wireless-enabled ICD were tested for potential interference with two models of recent 4G smartphones in conditions intended to maximize risk of EMI. A magnet effect (due to the phone speakers) was tested by placing the smartphones in the standby mode directly over the ICD generator. The presence of EMI artefacts on the real-time electrograms was evaluated by placing the smartphones in the standby, dialling, and operating modes directly over the generator casing and over the parasternal region in the vicinity of the ventricular lead. A total of 63 patients equipped with 29 different models of single, dual, or biventricular ICDs from five major manufacturers were included. None of the patients showed any evidence of interference with the smartphones during any of the 882 tests. The risk of EMI between modern smartphones and contemporary ICDs is low. This is probably due to the filters incorporated in the ICDs and low emission by the phones, as well as the small size of the magnets in the smartphones tested. NCT02330900 (http://www.clinicaltrials.gov). Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2016. For permissions please email: journals.permissions@oup.com.

  19. Implantable cardioverter-defibrillator explantation for overdiagnosed or overtreated congenital long QT syndrome.

    Science.gov (United States)

    Gaba, Prakriti; Bos, J Martijn; Cannon, Bryan C; Cha, Yong-Mei; Friedman, Paul A; Asirvatham, Samuel J; Ackerman, Michael J

    2016-04-01

    Primary treatment of long QT syndrome (LQTS) currently consists of beta-blocker therapy, although an implantable cardioverter-defibrillator (ICD) is considered for high-risk patients. However, both overdiagnosis and overtreatment must be avoided because their sequelae can be significant. The purpose of this study was to evaluate the prevalence and details of ICD explants in a cohort of patients from a tertiary genetic heart rhythm clinic for a previously rendered diagnosis of LQTS. Overall, 1227 consecutive patients were included in the study. All patients had been referred to the Mayo Clinic for evaluation of possible LQTS and subsequently were either diagnosed with LQTS or dismissed as normal. Further stratification of patients was conducted to assess how many patients had an ICD and how many had a subsequent ICD explant. In total, 170 patients (14%) had an ICD, including 157 of 670 patients (23%) with confirmed LQTS and 13 of 557 patients (2%) who did not have LQTS. Among these, 12 of 1227 (1%) had the ICD removed: 7 of 157 LQTS patients (4.5%) compared to 5 of 14 non-LQTS patients (36%). Before explant, 5 of 12 patients (42%) experienced inappropriate shocks, ranging from 2 to as many as 54 shocks. In addition, 4 had a device-related infection, and 9 had device malfunction (including lead dysfunction or fracture). None of these patients had a breakthrough cardiac event since removal of their ICD during 5.5 ± 3.5 years of follow-up. Implications of overdiagnosis and overtreatment are profound because unnecessary ICD placement can be associated with infection, malfunction, inappropriate shocks, and subsequent anxiety. Copyright © 2016 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

  20. Outcomes after asystole events occurring during wearable defibrillator-cardioverter use.

    Science.gov (United States)

    Liang, Jackson J; Bianco, Nicole R; Muser, Daniele; Enriquez, Andres; Santangeli, Pasquale; D'Souza, Benjamin A

    2018-04-26

    To examine whether wearable cardioverter defibrillator (WCD) alarms for asystole improve patient outcomes and survival. All asystole episodes recorded by the WCD in 2013 were retrospectively analyzed from a database of device and medical record documentation and customer call reports. Events were classified as asystole episodes if initial presenting arrhythmia was asystole (< 10 beats/minor ≥ 5 s pause). Survival was defined as recovery at the scene or arrival to a medical facility alive, or not requiring immediate medical attention. Episodes occurring in hospitals, nursing homes, or ambulances were considered to be under medical care. Serious asystole episodes were defined as resulting in unconsciousness, hospital transfer, or death. Of the total 51933 patients having worn the WCD in 2013, there were 257 patients (0.5%) who had asystole episodes and comprised the study cohort. Among the 257 patients (74% male, median age 69 years), there were 264 asystole episodes. Overall patient survival was 42%. Most asystoles were considered "serious" ( n = 201 in 201 patients, 76%), with a 26% survival rate. All 56 patients with "non-serious" asystole episodes survived. Being under medical care was associated with worse survival of serious asystoles. Among acute survivors, 20% later died during WCD use (a median 4 days post asystole episode). Of the 86 living patients at the end of WCD use period, 48 (56%) received ICD/pacemaker and 17 (20%) improved their condition. Survival rates after asystole in patients with WCD are higher than historically reported survival rates. Those under medical care at time of asystole exhibited lower survival.

  1. Feedback to providers improves evidence-based implantable cardioverter-defibrillator programming and reduces shocks.

    Science.gov (United States)

    Silver, Marc T; Sterns, Laurence D; Piccini, Jonathan P; Joung, Boyoung; Ching, Chi-Keong; Pickett, Robert A; Rabinovich, Rafael; Liu, Shufeng; Peterson, Brett J; Lexcen, Daniel R

    2015-03-01

    Implantable cardioverter-defibrillator (ICD) shocks are associated with increased anxiety, health care utilization, and potentially mortality. The purpose of the Shock-Less Study was to determine if providing feedback reports to physicians on their adherence to evidence-based shock reduction programming could improve their programming behavior and reduce shocks. Shock-Less enrolled primary prevention (PP) and secondary prevention (SP) ICD patients between 2009 and 2012 at 118 study centers worldwide and followed patients longitudinally after their ICD implant. Center-specific therapy programming reports (TPRs) were delivered to each center 9 to 12 months after their first enrollment. The reports detailed adherence to evidence-based programming targets: number of intervals to detect ventricular fibrillation (VF NID), longest treatment interval (LTI), supraventricular tachycardia (SVT) discriminators (Wavelet, PR Logic), SVT limit, Lead Integrity Alert (LIA), and antitachycardia pacing (ATP). Clinicians programmed ICDs at their discretion. The primary outcome measure was the change in utilization of evidence-based shock reduction programming before (phase I, n = 2694 patients) and after initiation of the TPR (phase II, n = 1438 patients). Patients implanted after feedback reports (phase II) were up to 20% more likely to have their ICDs programmed in line with evidence-based shock reduction programming (eg, VF NID in PP patients 30/40 in 33.5% vs 18.6%, P programming feedback reports improves adherence to evidence-based shock reduction programming and is associated with lower risk of ICD shocks. Copyright © 2015 Heart Rhythm Society. Published by Elsevier Inc. All rights reserved.

  2. 長野県内の病院におけるAED 普及状況

    OpenAIRE

    岩下, 具美; 二木, 智子; 関口, 幸男; 今村, 浩; 手塚, 理絵; 飯ケ濱, 実; 岡元, 和文

    2006-01-01

    Ventricular fibrillation (VF) is the main cause of sudden cardiac arrest. Because the mortality rate of patients with VF is extremely time-dependent, it is essential to perform defibrillation without delay. The automated external defibrillator (AED) is an instrument that laypeople can use anytime and anywhere. We surveyed the installation of AEDs in the 138 hospitals in Nagano Prefecture, and received 113 replies from the 138 hospitals. Fifty-nine hospitals (52%) have installed AEDs. Out of t...

  3. AutoDipole - Automated generation of dipole subtraction terms

    International Nuclear Information System (INIS)

    Hasegawa, K.; Uwer, P.

    2009-11-01

    We present an automated generation of the subtraction terms for next-to-leading order QCD calculations in the Catani-Seymour dipole formalism. For a given scattering process with n external particles our Mathematica package generates all dipole terms, allowing for bothmassless and massive dipoles. The numerical evaluation of the subtraction terms proceeds with MadGraph, which provides Fortran code for the necessary scattering amplitudes. Checks of the numerical stability are discussed. (orig.)

  4. AutoDipole - Automated generation of dipole subtraction terms

    Energy Technology Data Exchange (ETDEWEB)

    Hasegawa, K.; Uwer, P. [Humboldt-Universitaet, Berlin (Germany). Inst. fuer Physik; Moch, S. [Deutsches Elektronen-Synchrotron (DESY), Zeuthen (Germany)

    2009-11-15

    We present an automated generation of the subtraction terms for next-to-leading order QCD calculations in the Catani-Seymour dipole formalism. For a given scattering process with n external particles our Mathematica package generates all dipole terms, allowing for bothmassless and massive dipoles. The numerical evaluation of the subtraction terms proceeds with MadGraph, which provides Fortran code for the necessary scattering amplitudes. Checks of the numerical stability are discussed. (orig.)

  5. EO Model for Tacit Knowledge Externalization in Socio-Technical Enterprises

    Directory of Open Access Journals (Sweden)

    Shreyas Suresh Rao

    2017-03-01

    Full Text Available Aim/Purpose: A vital business activity within socio-technical enterprises is tacit knowledge externalization, which elicits and explicates tacit knowledge of enterprise employees as external knowledge. The aim of this paper is to integrate diverse aspects of externalization through the Enterprise Ontology model. Background: Across two decades, researchers have explored various aspects of tacit knowledge externalization. However, from the existing works, it is revealed that there is no uniform representation of the externalization process, which has resulted in divergent and contradictory interpretations across the literature. Methodology\t: The Enterprise Ontology model is constructed step-wise through the conceptual and measurement views. While the conceptual view encompasses three patterns that model the externalization process, the measurement view employs certainty-factor model to empirically measure the outcome of the externalization process. Contribution: The paper contributes towards knowledge management literature in two ways. The first contribution is the Enterprise Ontology model that integrates diverse aspects of externalization. The second contribution is a Web application that validates the model through a case study in banking. Findings: The findings show that the Enterprise Ontology model and the patterns are pragmatic in externalizing the tacit knowledge of experts in a problem-solving scenario within a banking enterprise. Recommendations for Practitioners\t: Consider the diverse aspects (what, where, when, why, and how during the tacit knowledge externalization process. Future Research:\tTo extend the Enterprise Ontology model to include externalization from partially automated enterprise systems.

  6. Classic conditioning and dysfunctional cognitions in patients with panic disorder and agoraphobia treated with an implantable cardioverter/defibrillator.

    Science.gov (United States)

    Godemann, F; Ahrens, B; Behrens, S; Berthold, R; Gandor, C; Lampe, F; Linden, M

    2001-01-01

    A model for the development of anxiety disorders (panic disorder with or without agoraphobia) is needed. Patients with an implantable cardioverter/defibrillator (ICD) are exposed to repeated electric shocks. If the theory of anxiety development by aversive classic conditioning processes is valid, these repeated shocks should lead to an increased risk of anxiety disorders. To study this hypothesis, we retrospectively studied 72 patients after implantation of an automatic ICD. Patients were assessed with the semistructured Diagnostic Interview of Psychiatric Disease 1 to 6 years after implantation of an automatic ICD. Panic disorder and/or agoraphobia was diagnosed in patients who fulfilled all DSM-III-R criteria for those conditions. Anxiety disorder developed in 15.9% of patients after ICD implantation. This was significantly related to the frequency of repeated defibrillation (shocks) to stop malignant ventricular arrhythmias. Dysfunctional cognitions are an additional vulnerability factor. The data support both the conditioning hypothesis and the cognitive model of anxiety development. These findings suggest that ICD patients are an appropriate risk population for a prospective study of the development of anxiety disorders.

  7. Canadian Registry of ICD Implant Testing procedures (CREDIT): current practice, risks, and costs of intraoperative defibrillation testing.

    Science.gov (United States)

    Healey, Jeff S; Dorian, Paul; Mitchell, L Brent; Talajic, Mario; Philippon, Francois; Simpson, Chris; Yee, Raymond; Morillo, Carlos A; Lamy, Andre; Basta, Magdy; Birnie, David H; Wang, Xiaoyin; Nair, Girish M; Crystal, Eugene; Kerr, Charles R; Connolly, Stuart J

    2010-02-01

    There is uncertainty about the proper role of defibrillation testing (DT) at the time of implantable cardioverter defibrillator (ICD) insertion. A prospective registry was conducted at 13 sites in Canada between January 2006 and October 2007. To document the details of DT, the reasons for not conducting DT, and the costs and complications associated with DT. DT was conducted at implantation in 230 of 361 patients (64%). DT was more likely to be conducted for new implants compared with impulse generator replacements (71% vs 32%, P = 0.0001), but was similar for primary and secondary prevention indications (64% vs 63%, P = NS). Among patients not having DT, the reason(s) given were: considered unnecessary (44%); considered unsafe, mainly due to persistent atrial fibrillation (37%); lack of an anesthetist (20%); and, patient or physician preference (6%). When performed, DT consisted of a single successful shock > or = 10J below maximum device output in 65% of cases. A 10J safety-margin was met by 97% of patients, requiring system modification in 2.3%. Major perioperative complications occurred in 4.4% of patients having DT versus 6.6% of patients not having DT (P = NS). ICD insertion was $844 more expensive for patients having DT (P = 0.16), largely due to increased costs ($28,017 vs $24,545) among patients having impulse generator replacement (P = 0.02). DT was not performed in a third of ICD implants, usually due to a perceived lack of need or relative contraindication.

  8. Improved extraction of ePTFE and medical adhesive modified defibrillation leads from the coronary sinus and great cardiac vein.

    Science.gov (United States)

    Wilkoff, Bruce L; Belott, Peter H; Love, Charles J; Scheiner, Avram; Westlund, Randy; Rippy, Marian; Krishnan, Mohan; Norlander, Barry E; Steinhaus, Bruce; Emmanuel, Janson; Zeller, Peter J

    2005-03-01

    Permanent leads with shocking coils for defibrillation therapy are sometimes implanted in the coronary sinus (CS) and great cardiac vein (GCV). These shocking coils, as documented by pathologic examination of animal investigations, often become tightly encapsulated by fibrosis and can be very difficult to remove. One of three configurations of the Guidant model 7109 Perimeter coronary sinus shocking lead was implanted into the distal portion of the GCV of 24 sheep for up to 14 months. Group 1 had unmodified coils (control), group 2 had coils backfilled with medical adhesive (MA), and Group 3 had coils coated with expanded polytetrafluoroethylene (ePTFE). Eighteen leads, three from each group at 6 and 14 months were transvenously extracted from the left jugular vein. The remaining six animals were not subject to extraction. All animals were euthanized for pathological and microscopic examination. All six of the control, three of the MA, and one of the ePTFE leads required the use of an electrosurgical dissection sheath (EDS) for extraction. Five control, two MA, and none of the ePTFE leads had significant fibrotic attachments to the shocking coils. Significant trauma was observed at necropsy for those leads requiring the use of the EDS for extraction. Tissue ingrowth is a major impediment to the removal of defibrillation leads implanted in the CS and GCV of sheep. Reduction of tissue ingrowth by coating the shocking coils with ePTFE or by backfilling with MA facilitates transvenous lead removal with reduced tissue trauma.

  9. Rapid automated nuclear chemistry

    International Nuclear Information System (INIS)

    Meyer, R.A.

    1979-01-01

    Rapid Automated Nuclear Chemistry (RANC) can be thought of as the Z-separation of Neutron-rich Isotopes by Automated Methods. The range of RANC studies of fission and its products is large. In a sense, the studies can be categorized into various energy ranges from the highest where the fission process and particle emission are considered, to low energies where nuclear dynamics are being explored. This paper presents a table which gives examples of current research using RANC on fission and fission products. The remainder of this text is divided into three parts. The first contains a discussion of the chemical methods available for the fission product elements, the second describes the major techniques, and in the last section, examples of recent results are discussed as illustrations of the use of RANC

  10. Automated optical assembly

    Science.gov (United States)

    Bala, John L.

    1995-08-01

    Automation and polymer science represent fundamental new technologies which can be directed toward realizing the goal of establishing a domestic, world-class, commercial optics business. Use of innovative optical designs using precision polymer optics will enable the US to play a vital role in the next generation of commercial optical products. The increased cost savings inherent in the utilization of optical-grade polymers outweighs almost every advantage of using glass for high volume situations. Optical designers must gain experience with combined refractive/diffractive designs and broaden their knowledge base regarding polymer technology beyond a cursory intellectual exercise. Implementation of a fully automated assembly system, combined with utilization of polymer optics, constitutes the type of integrated manufacturing process which will enable the US to successfully compete with the low-cost labor employed in the Far East, as well as to produce an equivalent product.

  11. Automated breeder fuel fabrication

    International Nuclear Information System (INIS)

    Goldmann, L.H.; Frederickson, J.R.

    1983-01-01

    The objective of the Secure Automated Fabrication (SAF) Project is to develop remotely operated equipment for the processing and manufacturing of breeder reactor fuel pins. The SAF line will be installed in the Fuels and Materials Examination Facility (FMEF). The FMEF is presently under construction at the Department of Energy's (DOE) Hanford site near Richland, Washington, and is operated by the Westinghouse Hanford Company (WHC). The fabrication and support systems of the SAF line are designed for computer-controlled operation from a centralized control room. Remote and automated fuel fabriction operations will result in: reduced radiation exposure to workers; enhanced safeguards; improved product quality; near real-time accountability, and increased productivity. The present schedule calls for installation of SAF line equipment in the FMEF beginning in 1984, with qualifying runs starting in 1986 and production commencing in 1987. 5 figures

  12. Automated multiple failure FMEA

    International Nuclear Information System (INIS)

    Price, C.J.; Taylor, N.S.

    2002-01-01

    Failure mode and effects analysis (FMEA) is typically performed by a team of engineers working together. In general, they will only consider single point failures in a system. Consideration of all possible combinations of failures is impractical for all but the simplest example systems. Even if the task of producing the FMEA report for the full multiple failure scenario were automated, it would still be impractical for the engineers to read, understand and act on all of the results. This paper shows how approximate failure rates for components can be used to select the most likely combinations of failures for automated investigation using simulation. The important information can be automatically identified from the resulting report, making it practical for engineers to study and act on the results. The strategy described in the paper has been applied to a range of electrical subsystems, and the results have confirmed that the strategy described here works well for realistically complex systems

  13. Automated drawing generation system

    International Nuclear Information System (INIS)

    Yoshinaga, Toshiaki; Kawahata, Junichi; Yoshida, Naoto; Ono, Satoru

    1991-01-01

    Since automated CAD drawing generation systems still require human intervention, improvements were focussed on an interactive processing section (data input and correcting operation) which necessitates a vast amount of work. As a result, human intervention was eliminated, the original objective of a computerized system. This is the first step taken towards complete automation. The effects of development and commercialization of the system are as described below. (1) The interactive processing time required for generating drawings was improved. It was determined that introduction of the CAD system has reduced the time required for generating drawings. (2) The difference in skills between workers preparing drawings has been eliminated and the quality of drawings has been made uniform. (3) The extent of knowledge and experience demanded of workers has been reduced. (author)

  14. ATLAS Distributed Computing Automation

    CERN Document Server

    Schovancova, J; The ATLAS collaboration; Borrego, C; Campana, S; Di Girolamo, A; Elmsheuser, J; Hejbal, J; Kouba, T; Legger, F; Magradze, E; Medrano Llamas, R; Negri, G; Rinaldi, L; Sciacca, G; Serfon, C; Van Der Ster, D C

    2012-01-01

    The ATLAS Experiment benefits from computing resources distributed worldwide at more than 100 WLCG sites. The ATLAS Grid sites provide over 100k CPU job slots, over 100 PB of storage space on disk or tape. Monitoring of status of such a complex infrastructure is essential. The ATLAS Grid infrastructure is monitored 24/7 by two teams of shifters distributed world-wide, by the ATLAS Distributed Computing experts, and by site administrators. In this paper we summarize automation efforts performed within the ATLAS Distributed Computing team in order to reduce manpower costs and improve the reliability of the system. Different aspects of the automation process are described: from the ATLAS Grid site topology provided by the ATLAS Grid Information System, via automatic site testing by the HammerCloud, to automatic exclusion from production or analysis activities.

  15. Automated Analysis of Accountability

    DEFF Research Database (Denmark)

    Bruni, Alessandro; Giustolisi, Rosario; Schürmann, Carsten

    2017-01-01

    that the system can detect the misbehaving parties who caused that failure. Accountability is an intuitively stronger property than verifiability as the latter only rests on the possibility of detecting the failure of a goal. A plethora of accountability and verifiability definitions have been proposed...... in the literature. Those definitions are either very specific to the protocols in question, hence not applicable in other scenarios, or too general and widely applicable but requiring complicated and hard to follow manual proofs. In this paper, we advance formal definitions of verifiability and accountability...... that are amenable to automated verification. Our definitions are general enough to be applied to different classes of protocols and different automated security verification tools. Furthermore, we point out formally the relation between verifiability and accountability. We validate our definitions...

  16. Beta-blocker therapy is not associated with symptoms of depression and anxiety in patients receiving an implantable cardioverter-defibrillator

    NARCIS (Netherlands)

    M.T. Hoogwegt (Madelein); N. Kupper (Nina); D.A.M.J. Theuns (Dominic); L.J.L.M. Jordaens (Luc); S.S. Pedersen (Susanne)

    2012-01-01

    textabstractBeta-blockers are frequently prescribed to implantable cardioverter-defibrillator (ICD) patients. Beta-blocker therapy has been proposed to induce emotional distress such as depression and anxiety, but a paucity of studies has examined the relationship between beta-blockers and distress.

  17. Comorbidity burden is associated with poor psychological well-being and physical health status in patients with an implantable cardioverter-defibrillator

    NARCIS (Netherlands)

    Hoogwegt, M.T.; Kupper, N.; Jordaens, L.; Pedersen, S.S.; Theuns, D.A.M.J.

    2013-01-01

    Aims Comorbidity burden has been linked to survival in patients with an implantable cardioverter-defibrillator (ICD), but no study has examined the influence on psychological well-being and health status. We examined the relationship between comorbidity burden and anxiety, depression, and health

  18. Comorbidity burden is associated with poor psychological well-being and physical health status in patients with an implantable cardioverter-defibrillator

    DEFF Research Database (Denmark)

    Hoogwegt, Madelein T; Kupper, Nina; Jordaens, Luc

    2013-01-01

    Comorbidity burden has been linked to survival in patients with an implantable cardioverter-defibrillator (ICD), but no study has examined the influence on psychological well-being and health status. We examined the relationship between comorbidity burden and anxiety, depression, and health status...

  19. Atrial fibrillation detection and R-wave synchronization by Metrix implantable atrial defibrillator - Implications for long-term efficacy and safety

    NARCIS (Netherlands)

    Tse, HF; Lau, CP; Sra, JS; Crijns, HJGM; Edvardsson, N; Kacet, S; Wyse, DG

    1999-01-01

    Background-The long-term efficacy of atrial fibrillation (AF) detection and R-wave synchronization are critical safety requirements for the development of an implantable atrial defibrillator (LAD) for treatment of AF. Methods and Results The long-term efficacy of the Metrix IAD for AF detection and

  20. Automation and Mankind

    Science.gov (United States)

    1960-08-07

    limited by the cap- abilities of the human organism in the matter of control of its processes. In our time, the speeds of technological processes are...in many cases limited by conditions of control. The speed of human reaction is limited and therefore, at pre- sent, only processes of a relatively...forwiard, It can e foreseer thast automIation will comp~letely free Mans -Pn work unler conlitions’ of high texpemratures pressures,, anid nollutA-: or