WorldWideScience

Sample records for area monitoring dosimeter

  1. Area monitoring dosimeter program for the Pacific Northwest National Laboratory: Results for CY 1997

    International Nuclear Information System (INIS)

    Bivins, S.R.; Stoetzel, G.A.

    1998-07-01

    In January 1993, Pacific Northwest National Laboratory (PNNL) established an area monitoring dosimeter program in accordance with Article 514 of the US Department of Energy (DOE) Radiological Control Manual (RCM). The purpose of the program was to minimize the number of areas requiring issuance of personnel dosimeters and to demonstrate that doses outside Radiological Buffer Areas are negligible. In accordance with 10 CFR Part 835.402 (a) (1)--(3) and Article 511.1 of the RCM, personnel dosimetry shall be provided to (1) radiological workers who are likely to receive at least 100 mrem annually, and (2) declared pregnant workers, minors, and members of the public who are likely to receive at least 50 mrem annually. Program results for calendar years (CY) 1993--1996 confirmed that personnel dosimetry was not needed for individuals located in areas monitored by the program. A total of 93 area thermoluminescent dosimeters (TLDs) were placed in PNNL facilities during CY 1997. The TLDs were exchanged and analyzed quarterly. All routine area monitoring TLD results were less than 50 mrem annually after correcting for worker occupancy. The results support the conclusions that personnel dosimeters are not necessary for staff, declared pregnant workers, minors, or members of the public in these monitored areas

  2. Area Monitoring Dosimeter Program for the Pacific Northwest National Laboratory: Results for CY 1998

    International Nuclear Information System (INIS)

    GA Stoetzel; SR Bivins

    1999-01-01

    In January 1993, Pacific Northwest National Laboratory (PNNL) established an area monitoring dosimeter program in accordance with Article 514 of the Department of Energy (DOE) Radiological Control Manual (RCM). The purpose of the program was to minimize the number of areas requiring issuance of personnel dosimeters and to demonstrate that doses outside Radiological Buffer Areas are negligible. In accordance with 10 CFR Part 835.402 (a)(1)-(4) and Article 511.1 of the RCM, personnel dosimetry shall be provided to (1) radiological workers who are likely to receive at least 100 mrem annually and (2) declared pregnant workers, minors, and members of the public who are likely to receive at least 50 mrem annually. Program results for calendar years 1993-1997 confirmed that personnel dosimetry was not needed for individuals located in areas monitored by the program. A total of 97 area thermoluminescent dosimeters (TLDs) were placed in PNNL facilities during calendar year 1998. The TLDs were exchanged and analyzed quarterly. All routine area monitoring TLD results were less than 50 mrem annually after correcting for worker occupancy. The results support the conclusion that personnel dosimeters are not necessary for staff, declared pregnant workers, minors, or members of the public in these monitored areas

  3. Area Monitoring Dosimeter Program for the Pacific Northwest National Laboratory: Results for CY 2000

    International Nuclear Information System (INIS)

    Bivins, Steven R; Stoetzel, Gregory A

    2001-01-01

    Pacific Northwest National Laboratory (PNNL) established an area monitoring dosimeter program in accordance with Article 514 of the Department of Energy (DOE) Radiological Control Manual (RCM) in January 1993. This program is to minimize the number of areas requiring issuance of personnel dosimeters and to demonstrate that doses outside Radiological Buffer Areas are negligible. In accordance with 10 CFR Part 835.402 (a) (1)-(4) and Article 511.1 of the DOE Standard Radiological Control, personnel dosimetry shall be provided to (1) radiological workers who are likely to receive at least 100 mrem annually, and (2) declared pregnant workers, minors, and members of the public who are likely to receive at least 50 mrem annually. Program results for calendar years 1993-2000 confirm that personnel dosimetry is not needed for individuals located in areas monitored by the program

  4. Directional Radiation Dosimeter for Area and Environmental Monitoring

    International Nuclear Information System (INIS)

    Manzoli, J.E.; Campos, V.P.; Moura, E.S.

    2009-01-01

    It is presented a dosimeter that is able to measure the photon exposure and the direction from where the radiation came from. Preliminary measurements performed by this new directional radiation dosimeter demonstrate its application. This dosimeter consists of a small lead cube with thermoluminescent discs on each face, placed in well known coordinates. Only one dosimeter of this kind indicates the direction of the radiation beam, if it came from a unique position. This study was conducted inside the radiation room of a Cobalt-60 Gamma Irradiator and the dosimeter indicated the source position

  5. Environmental monitoring by CaSO4:Dy TL dosimeters

    International Nuclear Information System (INIS)

    Deme, S.; Szabo, P.P.

    1975-12-01

    The thermoluminescent dosimeters of high sensitivity are useful for monitoring the area near nuclear installations. CaSO 4 :Dy TL dosimeters have high sensitivity and low fading so that by means of them the dose from the background can be measured with an accuracy of 10-20%. An increase of 2 mR in the background can be observed and doses as high as 1000R can be registered with an accuracy of 5%. The measuring method and results are reported here. For two years these CaSO 4 :Dy dosimeters have been successfully used at the site of the Central Research Institute for Physics. (K.A.)

  6. Automating the personnel dosimeter monitoring program

    International Nuclear Information System (INIS)

    Compston, M.W.

    1982-12-01

    The personnel dosimetry monitoring program at the Portsmouth uranium enrichment facility has been improved by using thermoluminescent dosimetry to monitor for ionizing radiation exposure, and by automating most of the operations and all of the associated information handling. A thermoluminescent dosimeter (TLD) card, worn by personnel inside security badges, stores the energy of ionizing radiation. The dosimeters are changed-out periodically and are loaded 150 cards at a time into an automated reader-processor. The resulting data is recorded and filed into a useful form by computer programming developed for this purpose

  7. Work place monitoring in accelerator facilities using thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Ribeiro, M.S.; Sanches, M.P.; Osima, A.M.; Rodriguez, D.L.; Carvalho, R.N.; Somessari, R.N.

    1998-01-01

    The increase in the use of large amounts of energy and large particles accelerators in development or in industrial processes for the reticulation, polymerization and sterilization of cables and wires allowed to discover and monitor work places in facilities having particle accelerators at the Institute of Energy and Nuclear Inquiries Comissao National de Energy Nuclear. Measures previously taken by technicians in routine monitoring, show that radiation doses found in the beams tube and at the door of the accelerator area is high enough to require routine programs to monitor work places at the installation. That is why, fifteen thermoluminescent dosimeters (TLD) where placed in different points of the facility where doses must be measured along a three month period and at the same time readings must be taken from control dosimeters kept within a shielded container. The monitor had a small double layer with three pellets of TLD CaSO4 Dy inside of a route carrier adopted in routine workers dosimetry usually. Outcomes show that the radiological protection program must be implemented to ameliorate and guarantee safety procedures

  8. Qualification of the monitor Pug-7N like dosimeter for neutrons

    International Nuclear Information System (INIS)

    Benites R, J. L.; Vega C, H. R.; Murillo O, R.; Velazquez F, J. B.

    2011-10-01

    By means of an inter-comparison method, the monitor for neutrons Pug-7N was enabled like dosimeter for neutrons of two magnitudes: the environmental equivalent dose, H*(10), and the H equivalent dose. The monitor Pug-7N has a plastic detector of scintillation Pns-20 that can be used inside or outside of its polyethylene cylindrical moderator. This designed to detect the neutrons presence that is shown in ana logical form by means of a fast count. Although the instrument is useful to detect the neutrons presence its design it does not allow to estimate the dose. With the purpose of enabling it as dosimeter for neutrons, their response was compared with the response of the area monitor for neutrons Bert hold Lb 6411 and Eberline NRD model Asp-1. Under the same irradiation conditions the 3 instruments were exposed to a source of 241 AmBe of 3.7E(9) Bq (100 mCi) of activity whose spectrum and dosimetric magnitudes were determined with a spectrometric system of Bonner spheres with scintillator of 6 Lil(Eu) and the NSDUAZ code. Conversion factors of H*(10)/cpm and H/cpm were obtained for the two options of the monitor detector Pug-7N, with this procedure the monitor Pug-7N besides determining the presence of neutrons, it has been enabled for their use as dosimeter for neutrons. (Author)

  9. Evaluation of performance of electronic dosimeters for individual monitoring: tests in laboratory

    International Nuclear Information System (INIS)

    Garzon, W.J.; Khoury, H.J.; Barros, V.S.M. de; Medeiros, R.B.

    2015-01-01

    Electronic dosimeters based on direct ion storage technology are being widely used in many countries for individual monitoring in many applications of ionizing radiation. However, their use as routine dosimeter has been established in a few countries due to lack of accreditation or intercomparison programs. The objective of this study is to evaluate the performance of two direct íon storage dosimeters model available in the international market: the Miriom-Instadose-1 and RADOS DIS-1 to be eventually accepted for individual monitoring in Brazil. (author)

  10. Radiation area monitoring by wireless-communicating area monitor with surveillance camera

    International Nuclear Information System (INIS)

    Shimura, Mitsuo; Kobayashi, Hiromitsu; Kitahara, Hideki; Kobayashi, Hironobu; Okamoto, Shinji

    2004-01-01

    Aiming at a dose reduction and a work efficiency improvement for nuclear power plants that have high dose regions, we have developed our system of wireless-communicating Area Monitor with Surveillance Camera, and have performed an on-site test. Now we are implementing this Area Monitor with Surveillance Camera for a use as a TV camera in the controlled-area, which enables a personal computer to simultaneously display two or more dose values and site live images on the screen. For the radiation detector of this Area Monitor System, our wireless-communicating dosimeter is utilized. Image data are transmitted via a wireless Local Area Network (LAN). As a test result, image transmission of a maximum of 20 frames per second has been realized, which shows that this concept is a practical application. Remote-site monitoring also has been realized from an office desk located within the non-controlled area, adopting a Japan's wireless phone system, PHS (Personal Handy Phone) for the transmission interface. (author)

  11. Ceramic BeO exoelectron dosimeters for tritium and radon monitoring

    International Nuclear Information System (INIS)

    Gammage, R.B.

    1979-01-01

    An environmental monitoring device with BeO ceramic dosimeters can be used to measure 222 Rn in dwellings. Radon diffuses into a porous hemispheric chamber and the radon daughters are electrostatically collected on aluminized Mylar foil covering the BeO dosimeter that records the alpha activity. A 10:1 signal-to-background ratio results from a radon exposure of only pCi-h/l. This high sensitivity makes accurate radon measurement possible within one day, even at near background levels of a few tenths pCi/l. The BeO exoelectron dosimeter is also uniquely suited for monitoring occupational exposure to insoluble tritium gas. At one-fifth the maximum permissible concentration, exposure for 8 hours gives a 10:1 signal-to-background exoelectron response to the low energy beta rays. Compensation for any exoelectron response caused by photon radiation can be made by reading the thermoluminescence. The tritium exposure produces negligible thermoluminescence. Progress in these and other applications is now totally dependent on achieving reliability and long-term stability of the exoelectron dosimeter

  12. Provision of dosimeters by official monitoring services for eye lens dose estimation

    International Nuclear Information System (INIS)

    Engelhardt, J.; Martini, E.

    2013-01-01

    Recent epidemiological studies are implying that the radio sensitivity of the eye lens is much higher than supposed in the past. International recommendations and standards demand to lower down the annual limit of the eye lens organ dose to 20 mSv. Since about 10 years German monitoring services offer partial-body dosimeters fixed on the head or on glasses for monitoring the eye lens dose. These dosimeters are optimized to measure the (surface) personal dose equivalent H p (0,07) from 0,5 mSv up to 10 Sv, which clearly overestimate the organ dose of the eye lens. With special features like different calibrations partial-body dosimeters should be applicable for legal dosimetry to avoid the development of special H p (3) dosimeters. Accepting the right way for wearing these dosimeters it is important to get the right results. Practical experiences are shown with measuring results and the difficulties of rounding the exact measuring values to discrete dose steps. Closing this article we point to still missing legal basis and open questions regarding to type testing procedures. (orig.)

  13. Evaluation of environmental monitoring thermoluminescent dosimeter locations

    International Nuclear Information System (INIS)

    Kinnison, R.

    1992-12-01

    Geostatistics, particularly kriging, has been used to assess the adequacy of the existing NTS thermoluminescent dosimeter network for determination of environmental exposure levels. (Kriging is a linear estimation method that results in contour plots of both the pattern of the estimated gamma radiation over the area of measurements and also of the standard deviations of the estimated exposure levels.) Even though the network was not designed as an environmental monitoring network, ft adequately serves this function in the region of Pahute and Rainier Mesas.. The Yucca Flat network is adequate only if a reasonable definition of environmental exposure levels is required; R is not adequate for environmental monitoring in Yucca Flat if a coefficient of variation of 10 percent or less is chosen as the criterion for network design. A revision of the Yucca Flat network design should be based on a square grid pattern with nodes 5000 feet (about one mile) apart, if a 10 percent coefficient of variation criterion is adopted. There were insufficient data for southern and western sections of the NTS to perform the geostatistical analysis. A very significant finding was that a single network design cannot be used for the entire NTS, because different areas have different variograms. Before any design can be finalized, the NTS management must specify the exposure unit area and coefficient of variation that are to be used as design criteria

  14. Qualification of the monitor Pug-7N like dosimeter for neutrons; Habilitacion del monitor PUG-7N como dosimetro para neutrones

    Energy Technology Data Exchange (ETDEWEB)

    Benites R, J. L. [Centro Estatal de Cancerologia de Nayarit, Av. Enfermeria, Fracc. Fray Junipero Serra, 63000 Tepic, Nayarit (Mexico); Vega C, H. R.; Murillo O, R. [Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Calle Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas (Mexico); Velazquez F, J. B., E-mail: jlbenitesr@prodigy.net.mx [Universidad Autonoma de Nayarit, Postgrado CBAP, Carretera Tepic Compostela Km. 9, Xalisco, Nayarit (Mexico)

    2011-10-15

    By means of an inter-comparison method, the monitor for neutrons Pug-7N was enabled like dosimeter for neutrons of two magnitudes: the environmental equivalent dose, H*(10), and the H equivalent dose. The monitor Pug-7N has a plastic detector of scintillation Pns-20 that can be used inside or outside of its polyethylene cylindrical moderator. This designed to detect the neutrons presence that is shown in ana logical form by means of a fast count. Although the instrument is useful to detect the neutrons presence its design it does not allow to estimate the dose. With the purpose of enabling it as dosimeter for neutrons, their response was compared with the response of the area monitor for neutrons Bert hold Lb 6411 and Eberline NRD model Asp-1. Under the same irradiation conditions the 3 instruments were exposed to a source of {sup 241}AmBe of 3.7E(9) Bq (100 mCi) of activity whose spectrum and dosimetric magnitudes were determined with a spectrometric system of Bonner spheres with scintillator of {sup 6}Lil(Eu) and the NSDUAZ code. Conversion factors of H*(10)/cpm and H/cpm were obtained for the two options of the monitor detector Pug-7N, with this procedure the monitor Pug-7N besides determining the presence of neutrons, it has been enabled for their use as dosimeter for neutrons. (Author)

  15. A passive environmental 222Rn monitor based on the exoelectron dosimeter

    International Nuclear Information System (INIS)

    Gammage, R.B.; Cheka, J.S.; Gesell, T.F.

    1978-01-01

    A high efficiency for BeO ceramic exoelectron dosimeters is demonstrated in the integrated monitoring of Mylar radon at concentrations close to natural background levels. Electrostatic collection of radon daughters onto aluminized mylar foil covering a BeO disk is achieved inside a porous, hemispherical chamber of the type developed by Costa-Ribeiro et al. This application of the exoelectron dosimeter for radon monitoring inside dwellings is a particularly favorable one; the lack of excessively high humidities and the clean conditions inside the hemisphere favor the reliable performance of the exoelectron dosimeter. Radon concentration - exposure times of 3 pCi h/l, or more, can be measured with an accuracy of about +-25% when the temperature and relative humidity are fluctuating. This means that radon concentrations of a few tenths of a pCi/l can be measured using exposure times of only a day or two. (Auth.)

  16. Environmental radiation monitoring of nuclear sites by thermoluminescent dosimeters (TLD)

    International Nuclear Information System (INIS)

    Duftschmid, K.E.; Strachotinsky, Ch.

    1978-04-01

    The measurement of environmental radiation doses around nuclear facilities requires the detection of few mrem/year. The properties of the automatic TLD-system Harshaw Mod. 2271 for such measurements have been evaluated under practical conditions and optimized techniques derived. The automatic TLD-system is based on LiF dosimeter cards with two crystals providing gamma and beta dose values. Limit of detection defined as three standard deviations of residuel dose is 1,2 mR. Automatic readout combined with electronic data evaluation are especially useful for large monitoring networks. Practical intercomparisons of this dosimeter with bulb-type CaF 2 detectors have been performed showing good agreement of both detector. Although bulb-dosimeters proved to be extremely sensitive with a limit of detection at 0,012 mR which makes them very suitable for very short exposure times, the automatic LiF system is superior in regards of man power requirement if monthly monitoring periods are sufficient. The system has been tested in practice during two international intercomparisons performed by the US Department of Energy - Health and Safety Laboratory New York and the Physikalisch Technische Bundesanstalt Braunschweig, Germany, showing excellent agreement. Furthermore a routine monitoring network consisting of 12 measurement positions around the Research Center Seibersdorf has been operated with this technique since more than two years. (author)

  17. Alarm pocket dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Hiraki, H; Kitamura, S [Matsushita Electric Industrial Co. Ltd., Kadoma, Osaka (Japan)

    1975-04-01

    This instrument is a highly reliable pocket dosimeter which has been developed for personal monitoring use. The dosimeter generates an alarm sound when the exposure dose reaches a preset value. Using a tiny GM tube for a radiation detector and measuring the integrated dose by means of a digital counting method, this new pocket dosimeter has high accuracy and stability. Using a sealed alkali storage battery for the power supply, and with an automatic control charger, this dosimetry system is easy and economical to operate and maintain. Detectable radiation by the dosimeter are X and ..gamma.. rays. Standard preset dose values are 30, 50, 80 and 100 mR. Detection accuracy is betwen +10% and -20%. The dosimeter is continuously usable for more than 14 hours after charging for 2 hours. The dosimeter has the following features; good realiability, shock-proof loud and clear alarm sound, the battery charger also serves as a stock container for the dosimeters, and no switching operation required for the power supply due to the internal automatic switch. Therefore, the dosimetry system is very useful for personal monitoring management in many radiation industry establishments.

  18. NRC TLD [Nuclear Regulatory Commission thermoluminescent dosimeter] direct radiation monitoring network

    International Nuclear Information System (INIS)

    Struckmeyer, R.; McNamara, N.

    1990-03-01

    This report presents the results of the NRC Direct Radiation Monitoring Network for the fourth quarter of 1989. It provides the ambient radiation levels measured in the vicinity of 75 sites throughout the United States. In addition, it describes the equipment used, monitoring station selection criteria, characterization of the dosimeter response, calibration procedures, statistical methods, intercomparison, and quality assurance program

  19. NRC [Nuclear Regulatory Commission] TLD [thermoluminescent dosimeter] direct radiation monitoring network

    International Nuclear Information System (INIS)

    Struckmeyer, R.; McNamara, N.

    1989-09-01

    This report provides the status and results of the NRC Thermoluminescent Dosimeter (TLD) Direct Radiation Monitoring Network. It presents the radiation levels measured in the vicinity of NRC licensed facility sites throughout the country for the second quarter of 1989

  20. Pen dosimeters

    CERN Multimedia

    SC/RP Group

    2006-01-01

    The Radiation Protection Group has decided to withdraw all pen dosimeters from the main PS and SPS access points. This will be effective as of January 2006. The following changes will be implemented: All persons working in a limited-stay controlled radiation area must wear an operational dosimeter in addition to their personal DIS dosimeter. Any persons not equipped with this additional dosimeter must contact the SC/RP Group, which will make this type of dosimeter available for temporary loan. A notice giving the phone numbers of the SC/RP Group members to contact will be displayed at the former distribution points for the pen dosimeters. Thank you for your cooperation. The SC/RP Group

  1. Use of an albedo neutron personnel dosimeter for X- and γ-ray monitoring

    International Nuclear Information System (INIS)

    Gorbics, S.G.; Nash, A.E.; Johnson, T.L.

    1981-01-01

    With a judicious choice of cadmium filter size and thickness, it is possible to use the information from the 7 LiF detectors used in an albedo neutron personnel dosimeter to determine an individual's X-and γ-ray exposure, thus eliminating the need for a separate dosimeter for this purpose. A filter area of 400 mm 2 and a thickness of 0.51 mm is shown to be optimum for a simple, plastic, dosimeter design using detectors held in dental-film size cards. (author)

  2. NRC TLD [thermoluminescent dosimeter] Direct Radiation Monitoring Network: Progress report, October--December 1988

    International Nuclear Information System (INIS)

    Struckmeyer, R.; NcNamara, N.

    1989-04-01

    This report presents the results of the NRC Direct Radiation Monitoring Network for the fourth quarter of 1988. It provides the ambient radiation levels measured in the vicinity of 75 sites throughout the United States. In addition, it describes the equipment used, monitoring station selection criteria, characterization of the dosimeter response, calibration procedures, statistical methods, intercomparison, and quality assurance program. 4 tabs

  3. NRC TLD [thermoluminescent dosimeter] Direct Radiation Monitoring Network: Progress report, January-March 1988

    International Nuclear Information System (INIS)

    Struckmeyer, R.; McNamara, N.

    1988-06-01

    This report provides the status and results of the NRC Thermoluminescent Dosimeter (TLD) Direct Radiation Monitoring Network. It presents the radiation levels measured in the vicinity of NRC licensed facility sites throughout the country for the first quarter of 1988

  4. Development of bead-type radiophotoluminescence glass dosimeter applicable to various purposes

    International Nuclear Information System (INIS)

    Sato, F.; Toyota, Y.; Maki, D.; Zushi, N.; Kato, Y.; Yamamoto, T.; Iida, T.

    2013-01-01

    Bead-type radiophotoluminescence (RPL) glass dosimeters were well fabricated with a gas-particle jet flame system for glass melting-cooling process. A rod of silver-activated phosphate glass was pulverized into micrometer-size particles. Spherical glass particles were formed from the pulverized glass particles in the high-temperature jet flame owing to the surface tension of the glass material. Some groups of spherical glass particles were irradiated with X-rays and their RPL was demonstrably observed for their exposure to UV light. A flexible RPL glass sheet was also made of bead-type RPL glass dosimeters and was useful for radiation imaging. Bead-type RPL glass dosimeters are expected to be used for dose monitoring in highly radioactively-contaminated area. -- Highlights: ► We developed bead-type radiophotoluminescence glass dosimeters. ► Bead-type glass RPL dosimeters are satisfactorily used as radiation dosimeters. ► A flexible RPL glass sheet is made of bead-type RPL glass dosimeters

  5. Characteristics of radiophotoluminescent glass dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Ito, Masashi; Shiraishi, Akemi; Murakami, Hiroyuki [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    2001-07-01

    In Japan Atomic Energy Research Institute, a film badge is recently replaced by a new type radiophotoluminescent (RPL) glass dosimeter for external personal monitoring. Some fundamental characteristics of this dosimeter, such as dose dependence linearity, energy dependence, angular dependence, dose evaluation accuracy at mixed irradiation conditions, fading, etc., were examined at the Facility of Radiation Standard (FRS), JAERI. The results have proved that the RPL glass dosimeter has sufficient characteristics for practical use as a personal dosimeter for all of the items examined. (author)

  6. Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF:Mg,Cu,P thermoluminescence dosimeters for use in diagnostic radiology

    International Nuclear Information System (INIS)

    Dong, S.L.; Chu, T.C.; Lan, G.Y.; Wu, T.H.; Lin, Y.C.; Lee, J.S.

    2002-01-01

    Monitoring radiation exposure during diagnostic radiographic procedures has recently become an area of interest. In recent years, the LiF:Mg,Cu,P thermoluminescence dosimeter (TLD-100H) and the highly sensitive metal oxide semiconductor field effect transistor (MOSFET) dosimeter were introduced as good candidates for entrance skin dose measurements in diagnostic radiology. In the present study, the TLD-100H and the MOSFET dosimeters were evaluated for sensitivity, linearity, energy, angular dependence, and post-exposure response. Our results indicate that the TLD-100H dosimeter has excellent linearity within diagnostic energy ranges and its sensitivity variations were under 3% at tube potentials from 40 Vp to 125 kVp. Good linearity was also observed with the MOSFET dosimeter, but in low-dose regions the values are less reliable and were found to be a function of the tube potentials. Both dosimeters also presented predictable angular dependence in this study. Our findings suggest that the TLD-100H dosimeter is more appropriate for low-dose diagnostic procedures such as chest and skull projections. The MOSFET dosimeter system is valuable for entrance skin dose measurement with lumbar spine projections and certain fluoroscopic procedures

  7. Neutron area monitor with TLD pairs

    International Nuclear Information System (INIS)

    Guzman G, K. A.; Borja H, C. G.; Valero L, C.; Hernandez D, V. M.; Vega C, H. R.

    2011-11-01

    The response of a passive neutron area monitor with pairs of thermoluminescent dosimeters has been calculated using the Monte Carlo code MCNP5. The response was calculated for one TLD 600 located at the center of a polyethylene cylinder, as moderator. When neutrons collide with the moderator lose their energy reaching the TLD with thermal energies where the ambient dose equivalent is calculated. The response was calculated for 47 monoenergetic neutron sources ranging from 1E(-9) to 20 MeV. Response was calculated using two irradiation geometries, one with an upper source and another with a lateral source. For both irradiation schemes the response was calculated with the TLDs in two positions, one parallel to the source and another perpendicular to the source. The advantage of this passive neutron monitor area is that can be used in locations with intense, pulsed and mixed radiation fields. (Author)

  8. Comparing personal alpha dosimetry with the conventional area monitoring-time weighting methods of exposure estimation: a Canadian assessment

    International Nuclear Information System (INIS)

    Balint, A.B.; Viljoen, J.

    1988-01-01

    An experimental personal alpha dosimetry program for monitoring exposures of uranium mining facility workers in Canada has been completed. All licenced operating mining facilities were participating. Dosimetry techniques, description of dosimeters used by licences, performance and problems associated with the implementation of the programme as well as technical and administrative advantages and difficulties experienced are discussed. Area monitoring-time weighting methods used and results obtained to determine individual radon and thoron daughter exposure and exposure results generated by using dosimeters are assessed and compared

  9. Effect of electromagnetic field in fusion facility on electronic personal dosimeter

    International Nuclear Information System (INIS)

    Yamada, Junya; Kawano, Takao; Uda, Tatsuhiko; Shimo, Michikuni

    2010-01-01

    The effect of electromagnetic field on electronic personal dosimeters in a nuclear fusion facility was examined in a Magnetic Resonance Imaging (MRI) examination room instead of a nuclear fusion facility. Three types of electronic personal dosimeters, the PDM-111, the 112, and the 117, were used as typical ones. We surveyed the electromagnetic field distribution and dosimeters were placed at locations with various strengths of the electromagnetic field. The natural radiation dose was measured for about one week. We found that while dosimeters were not affected by the electric field, they were affected by the magnetic one. Dosimeters detected radiation levels less sensitively as the magnetic field strength was increased up to 150 mT. The dosimeters underestimated the environmental radiation dose rates by about 10-30% when the magnetic field strength was larger than 150 mT. We assumed that hall-effect caused the reduction in radiation sensitivity. We concluded that the strength of the magnetic field needs to be carefully considered when an electronic personal dosimeter is used for monitoring both personal and area dose in a nuclear fusion facility. (author)

  10. Cell-phone interference with pocket dosimeters

    International Nuclear Information System (INIS)

    Djajaputra, David; Nehru, Ramasamy; Bruch, Philip M; Ayyangar, Komanduri M; Raman, Natarajan V; Enke, Charles A

    2005-01-01

    Accurate reporting of personal dose is required by regulation for hospital personnel that work with radioactive material. Pocket dosimeters are commonly used for monitoring this personal dose. We show that operating a cell phone in the vicinity of a pocket dosimeter can introduce large and erroneous readings of the dosimeter. This note reports a systematic study of this electromagnetic interference. We found that simple practical measures are enough to mitigate this problem, such as increasing the distance between the cell phone and the dosimeter or shielding the dosimeter, while maintaining its sensitivity to ionizing radiation, by placing it inside a common anti-static bag. (note)

  11. Cell-phone interference with pocket dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Djajaputra, David; Nehru, Ramasamy; Bruch, Philip M; Ayyangar, Komanduri M; Raman, Natarajan V; Enke, Charles A [Department of Radiation Oncology, University of Nebraska Medical Center, 987521 Nebraska Medical Center, Omaha, NE 68198-7521 (United States)

    2005-05-07

    Accurate reporting of personal dose is required by regulation for hospital personnel that work with radioactive material. Pocket dosimeters are commonly used for monitoring this personal dose. We show that operating a cell phone in the vicinity of a pocket dosimeter can introduce large and erroneous readings of the dosimeter. This note reports a systematic study of this electromagnetic interference. We found that simple practical measures are enough to mitigate this problem, such as increasing the distance between the cell phone and the dosimeter or shielding the dosimeter, while maintaining its sensitivity to ionizing radiation, by placing it inside a common anti-static bag. (note)

  12. Evaluation of personal dosimeters

    International Nuclear Information System (INIS)

    Correa, C. A.

    2007-01-01

    This work makes a screening of the different types of dosimeters present in the international market, to provide operative dosimetry of individual monitoring to measure Hp(10) and Hp(0,07)-specifically for external radiation gamma and beta, as well as to give knowledge of advances of passive and operative dosimetry, and the changes in the regulatory policy relative to these aspects. The data has been extracted from several providers of dosimeters, and the importance has been stressed in a good election of the dosimeter before its use, as well as the important advances in these equipment. (Author) 14 refs

  13. Offsite environmental monitoring report: Radiation monitoring around United States nuclear test areas

    International Nuclear Information System (INIS)

    1990-01-01

    This report describes the Offsite Radiation Safety Program conducted during 1990 by the Environmental Protection Agency's (EPA's) Environmental Monitoring Systems Laboratory -- Las Vegas. This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ion chambers (PICs); and by biological monitoring of animals, food crops, and humans. Personnel with mobile monitoring equipment are placed in areas downwind from the test site prior to each nuclear weapons test to implement protective actions, provide immediate radiation monitoring, and obtain environmental samples rapidly after any occurrence of radioactivity release

  14. Calibration of radiation protection area monitoring instruments in Sudan

    International Nuclear Information System (INIS)

    Suliman, I.I.; Youssif, B.E.; Beineen, A.A.; Hassan, M.

    2010-01-01

    This article presents results of measurements for the calibration of radiation protection area monitoring instruments carried out during the period 2006-2008 at Secondary Standard Dosimetry Laboratory of Sudan. The work performed included quality assurance measurements, measurements for the dosimeter calibrations and uncertainty analysis. Calibrations were performed using 137 Cs gamma ray sources produced by OB 85 and OB 34/1 gamma calibrators producing air kerma rate that ranged from 10 μGy/h to 50 mGy/h. Area monitoring instruments were calibrated in terms of ambient dose equivalent, H*(10) derived using air kerma to ambient dose equivalent conversion coefficients. Results are presented for 78 area monitoring instruments representing most commonly used types in Sudan. Radioactive check source measurements for the reference chamber showed deviation within 1% limit. The accuracy in the beam output measurements was within 5% internationally considered as acceptable. The results highlighted the importance of radiation protection calibrations. Regulations are further need to ensure safety aspect really meet the required international standards.

  15. Radiation monitoring at Pakistan research reactor

    International Nuclear Information System (INIS)

    Ali, A.

    1984-05-01

    Area radiation monitoring is accomplished by using Tracer Lab. radiation monitor. Personnel monitoring is carried out using film badges, TLDs (Thermoluminescent Dosimeters) and pocket dosimeters. For the evaluation of monthly accumulated doses of radiation workers film badges/TLDs and for instantaneous/short term dose measurement in higher radiation zones pocket dosimeters are used in addition to film badge/TLD. Environmental monitoring is necessary to check the PARR operation effect on background radiation level in the vicinity of PINSTECH. (A.B.). 4 refs

  16. Evaluation of performance of electronic dosimeters for individual monitoring: tests in laboratory; Avaliacao do desempenho de dosimetros eletronicos para monitoracao individual: testes em laboratorio

    Energy Technology Data Exchange (ETDEWEB)

    Garzon, W.J.; Khoury, H.J.; Barros, V.S.M. de, E-mail: wjaramig@gmail.com [Universidade Federal de Pernambuco (UFPE), Recife (Brazil). Departamento de Energia Nuclear. Laboratorio de Metrologia das Radiacoes Ionizantes; Medeiros, R.B. [Universidade Federal de Sao Paulo (EPM/UNIFESP), Sao Paulo, SP (Brazil). Departamento de Diagnostico por Imagem

    2015-07-01

    Electronic dosimeters based on direct ion storage technology are being widely used in many countries for individual monitoring in many applications of ionizing radiation. However, their use as routine dosimeter has been established in a few countries due to lack of accreditation or intercomparison programs. The objective of this study is to evaluate the performance of two direct íon storage dosimeters model available in the international market: the Miriom-Instadose-1 and RADOS DIS-1 to be eventually accepted for individual monitoring in Brazil. (author)

  17. Comparison of electronic digital alarm dosimeter with TLD

    International Nuclear Information System (INIS)

    Kumar, Pankaj; Pandey, J.P.N.; Shinde, A M.; Purohit, R.G.; Sarkar, P.K.

    2012-01-01

    Control of exposure of radiation workers on day to day basis has been made easy by use of semiconductor based electronic digital dosimeter. Additional dose constraints of 10 mSv for occupational radiation workers have made it essential to use such type of digital personal monitoring devices. In addition to conventional ionisation chamber based direct reading dosimeters, additional 35 semiconductor based digital dosimeters model MGP DMC 2000 S were used for the monitoring of personal exposure of radiation workers in a spent fuel reprocessing plant. Though better least count and good performance over a wide range of dose rate are claimed by the manufacture, before making use of such dosimeter on large scale, validation of its performance is required to be checked. In this paper, an effort is made to determine the performance of digital dosimeters, by exposing these digital dosimeters in combination with TLDs at different radiation levels and obtained results were compared and analysed

  18. Tracking and Monitoring with Dosimeter-Enabled ARG-US RFID System - 12009

    Energy Technology Data Exchange (ETDEWEB)

    Anderson, J.; Lee, H.; De Lurgio, P.; Kearney, C.M.; Craig, B.; Soos, I.H.; Tsai, H.; Liu, Y. [Argonne National Laboratory, Argonne, IL 60439 (United States); Shuler, J. [U.S. Department of Energy, Washington, D.C. 20585 (United States)

    2012-07-01

    Automated monitoring and tracking of materials with radio frequency identification (RFID) technology can significantly improve both the operating efficiency of radiological facilities and the application of the ALARA (as low as reasonably achievable) principle in them. One such system, called ARG-US, has been developed by Argonne National Laboratory for the U.S. Department of Energy (DOE) Packaging and Certification Program to use in managing sensitive nuclear and radioactive materials. Several ARG-US systems are in various stages of deployment and advanced testing across DOE sites. ARG-US utilizes sensors in the tags to continuously monitor the state of health of the packaging and promptly disseminates alarms to authorized users. In conjunction with global positioning system (GPS) tracking provided by TRANSCOM, the system can also monitor and track packages during transport. A compact dosimeter has been incorporated in the ARG-US tags via an onboard universal asynchronous receiver/transmitter interface. The detector has a wide measurement range for gamma radiation - from 0.1 mSv/h to 8 Sv/h. The detector is able to generate alarms for both high and low radiation and for a high cumulative dose. In a large installation, strategically located dosimeter-enabled tags can yield an accurate, real-time, 2D or 3D dose field map that can be used to enhance facility safety, security, and safeguards. This implementation can also lead to a reduced need for manned surveillance and reduced exposure of personnel to radiation, consistent with the ALARA principle at workplaces. (authors)

  19. Dose measurement during defectoscopic work using electronic personal dosimeters

    International Nuclear Information System (INIS)

    Smoldasova, J.

    2008-01-01

    Personal monitoring of the external radiation of radiation, personnel exposed to sources of ionizing radiation at a workplace is an important task of the radiological protection. Information based on the measured quantities characterizing the level of the exposure of radiation personnel enable to assess the optimum radiological protection at the relevant workplace and ascertain any deviation from the normal operation in time. Different types of personal dosimeters are used to monitor the external radiation of radiation personnel. Basically, there are two types of dosimeters, passive and active (electronic). Passive dosimeters provide information on the dose of exposure after its evaluation, while electronic dosimeters provide this information instantly. The goal of the work is to compare data acquired during different working activities using the DMC 2000 XB electronic dosimeters and the passive film dosimeters currently used at the defectoscopic workplace. (authors)

  20. The radiation monitoring system for the LHC experiments and experimental areas

    CERN Document Server

    Ilgner, C

    2004-01-01

    With the high energies stored in the beams of the LHC, special attention needs to be paid to accident scenarios involving beam losses which may have an impact on the installed experiments. Among others, an unsynchronized beam abort and a D1 magnet failure are considered serious cases. According to simulations, the CMS inner tracker in such accident scenarios can be damaged by instantaneous rates which are many orders of magnitude above normal conditions. Investigations of synthetic diamond as a beam condition monitor sensor, capable of generating a fast beam dump signal, will be presented. Furthermore, a system to monitor the radiation fields in the experimental areas is being developed. It must function in the radiation fields inside and around the experiments, over a large dynamic range. Several new active and passive sensors, such as RadFET, OSL (Optically Stimulated Luminescence) sensors, p-i-n diodes, Polymer-Alanine Dosimeters and TLDs (Thermoluminescent Dosimeters) are under investigation. Recent resul...

  1. Radiation dosimeter assembly

    International Nuclear Information System (INIS)

    Seidel, J.G.

    1982-01-01

    A technique is disclosed for securing a thermoluminescent radiation dosimeter, used for monitoring underground radon gas in uranium prospecting, to a cup-like support member made of heavy gauge aluminum foil. A metalized film, consisting of an aluminum layer and a high tensile strength plastic layer, covers an aperture in the support members for the dosimeter. The film is secured by a high temperature adhesive to the support member, and both are capable of withstanding an annealing temperature of up to 300 0 C

  2. To the attention of all dosimeter users

    CERN Multimedia

    2005-01-01

    Many regular users of CERN personal dosimeters do not respect the safety regulations, which include a compulsory monthly read-out of the dosimeter. Therefore we would like to remind everybody that if the dosimeter is not read for a period of 3 months or more, we will ask for a return or replacement of the dosimeter, which has a value of CHF 350.-. We would like to emphasise that the dosimeter must be read even if you have not entered controlled areas. Staff members or CERN users who enter controlled areas only occasionally may exchange their regular dosimeter for a short term visitor dosimeter (VCT). This dosimeter has a limited validity period but without for a compulsory periodic read-out. For further information please contact dosimetry.service@cern.ch Thank you for your cooperation. Dosimetry Service Bldg. 24 E 011 http://cern.ch/rp-dosimetry

  3. To the attention of all dosimeter users

    CERN Multimedia

    Dosimetry Service

    2005-01-01

    Many regular users of CERN personal dosimeters do not respect the safety regulations, which include the compulsory monthly read-out of the dosimeter. Therefore we would like to remind everybody that if the dosimeter is not read for a period of 3 months or more, we will ask for a return or replacement of the dosimeter, which has a value of CHF 350.-. We would like to emphasise that the dosimeter must be read even if you have not entered controlled areas. Staff members or CERN users who enter controlled areas only occasionally may exchange their regular dosimeter for a short term visitor dosimeter (VCT). This dosimeter has a limited validity period but without for a compulsory periodic read-out. For further information please contact dosimetry.service@cern.ch Thank you for your cooperation. Dosimetry Service Bld 24 E 011 http://cern.ch/rp-dosimetry

  4. To the attention of all dosimeter users

    CERN Multimedia

    Dosimetry Service

    2006-01-01

    Many regular users of CERN personal dosimeters do not respect the safety regulations, which include the compulsory monthly read-out of the dosimeter. We would therefore like to remind everybody that if the dosimeter is not read for a period of 3 months or more, we will ask for a return or replacement of the dosimeter, which has a value of CHF 350,-. We would like to emphasise that the dosimeter must be read even if you have not entered controlled areas. Staff members or CERN users who enter controlled areas only occasionally may exchange their regular dosimeter for a short-term visitor dosimeter (VCT). This dosimeter has a limited validity period but does not require a periodic read-out. For further information please contact dosimetry.service@cern.ch Thank you for your cooperation. Dosimetry Service - Bldg. 24 E 011 - http://cern.ch/rp-dosimetry

  5. Composite Resin Dosimeters: A New Concept and Design for a Fibrous Color Dosimeter.

    Science.gov (United States)

    Kinashi, Kenji; Iwata, Takato; Tsuchida, Hayato; Sakai, Wataru; Tsutsumi, Naoto

    2018-04-11

    Polystyrene (PS)-based composite microfibers combined with a photochromic spiropyran dye, 1,3,3-trimethylindolino-6'-nitrobenzopyrylospiran (6-nitro BIPS), and a photostimulable phosphor, europium-doped barium fluorochloride (BaFCl:Eu 2+ ), were developed for the detection of X-ray exposure doses on the order of approximately 1 Gy. To produce the PS-based composite microfibers, we employed a forcespinning method that embeds a high concentration of phosphor in PS in a safe, inexpensive, and simple procedure. On the basis of the optimization of the forcespinning process, fibrous color dosimeters with a high radiation dose sensitivity of 1.2-4.4 Gy were fabricated. The color of the dosimeters was found to transition from white to blue in response to X-ray exposure. The optimized fibrous color dosimeter, made from a solution having a PS/6-nitro BIPS/BaFCl:Eu 2+ /C 2 Cl 4 ratio of 7.0/0.21/28.0/28.0 (wt %) and produced with a 290 mm distance between the needle and collectors, a 0.34 mm 23 G needle nozzle, and a spinneret rotational rate of 3000 rpm, exhibited sensitivity to a dose as low as 1.2 Gy. To realize practical applications, we manufactured the optimized fibrous color dosimeter into a clothlike color dosimeter. The clothlike color dosimeter was mounted on a stuffed bear, and its coloring behavior was demonstrated upon X-ray exposure. After exposure with X-ray, a blue colored and shaped in the form of the letter "[Formula: see text]" clearly appeared on the surface of the clothlike color dosimeter. The proposed fibrous color dosimeters having excellent workability will be an unprecedented dosimetry and contributed to all industries utilizing radiation dosimeters. This new fibrous "composite resin dosimeter" should be able to replace traditional, wearable, and individual radiation dose monitoring devices, such as film badges.

  6. Mayak Film Dosimeter Response Studies Part III: Application to Worker Dose Assessment

    International Nuclear Information System (INIS)

    Smetanin, Mikhail; Vasilenko, E. K.; Scherpelz, Robert I.

    2007-01-01

    This paper describes the methods used to convert individual dosimeter readings for workers to obtain estimates of worker doses received in Mayak facilities. Film dosimeters were used at Mayak PA for worker monitoring from 1948 until 1992. The method requires a determination of the relationship between the absorbed dose in film emulsion and the dose in air under calibration conditions, then an extension of this relationship to exposures in the actual radiation fields of the workplace. Corrections needed to account for actual workplace exposure conditions were determined by modeling with the Monte-Carlo radiation transport computer code MCNP. Correction factors were developed to convert from dosimeter reading to a realistic worker dose. The method was applied as a basis for individual dose reconstruction using film dosimeters in realistic photon spectra and geometries at Mayak PA work areas

  7. A new radiochromic dosimeter film

    Science.gov (United States)

    Sidney, L. N.; Lynch, D. C.; Willet, P. S.

    By employing acid-sensitive leuco dyes in a chlorine-containing polymer matrix, a new radiochromic dosimeter film has been developed for gamma, electron beam, and ultraviolet radiation. These dosimeter films undergo a color change from colorless to royal blue, red fuchsia, or black, depending on dye selection, and have been characterized using a visible spectrophotometer over an absorbed dose range of 1 to 100 kGy. The primary features of the film are improved color stability before and after irradiation, whether stored in the dark or under artificial lights, and improved moisture resistance. The effects of absorbed dose, dose rate, and storage conditions on dosimeter performance are discussed. The dosimeter material may be produced as a free film or coated onto a transparent substrate and optionally backed with adhesive. Potential applications for these materials include gamma sterilization indicator films for food and medical products, electron beam dosimeters, and in-line radiation monitors for electron beam and ultraviolet processing.

  8. Copper doped borate dosimeters revisited

    International Nuclear Information System (INIS)

    Alajerami, Y.S.M.; Hashim, S.; Ghoshal, S.K.; Bradley, D.A.; Mhareb, M.; Saleh, M.A.

    2014-01-01

    We render a panoramic overview on copper (Cu) doped borate dosimeters. Preparing a dosimeter by mixing specific materials with precise weights and methods is a never-ending quest. The recommended composition is highly decisive for accurate estimation of the absorbed dose, prediction of the biological outcome, determination of the treatment dose for radiation therapy and facilitation of personal monitoring. Based on these principles, the proposed dosimeter must cover a series of dosimetric properties to realize the exact results and assessment. The doped borate dosimeters indeed demonstrate attractive thermoluminescence (TL) features. Several dedicated efforts are attempted to improve the luminescence properties by doping various transition metals or rare-earth elements. The Cu ion being one of the preferred activators shows excellent TL properties as revealed via detail comparison with other dosimeters. Two oxide states of Cu (Cu + and Cu ++ ) with reasonable atomic number allow easy interaction with boron network. Interestingly, the intrinsic luminescent centers of borate lattice are in cross linked with that of Cu + ions. Thus, the activation of borate dosimeter with Cu ions for the enhancement of the TL sensitivity is recognized. These dosimeters reveal similar glow curves as the standard TLD-100 (LiF:Mg,Ti) one irrespective of the use of modifiers and synthesis techniques. They display high sensitivity, low fading, dose response linearity over wide range and practical minimum detectable dose. Furthermore, the effective atomic number being the most beneficial aspect (equivalent to that of human tissue) of borate dosimeters do not show any change due to Cu ion activations. The past development, major challenges, excitement, applications, recent progress and the future promises of Cu doped borate TL dosimeters are highlighted. - Highlights: • The manuscript gives a panoramic overview on copper doped borate dosimeters. • Cu ions activated technique in borate

  9. The New HARSHAW Extremity Dosimeters for Gamma and Beta Ray Monitoring

    International Nuclear Information System (INIS)

    Fellinger, J.; Majewski, M.; Rotunda, J.; Tawi, R.

    1997-01-01

    Large personnel dosimetry services providing extremity monitoring with finger rings based on thermoluminescent detectors have long been looking for a practical method for automated reading including automated identification of the detectors.All existing methods are at least not very suitable for medical applications, particularly for surgery, due to the fact that cold sterilization is usually impossible.Bicron radiation Measurement Products developed in co-operation with the Austrian Research Centre Seibersdorf a new finger ring dosimeter DXT-RAD as a fast and economic solution for fully automated evaluation of extremity dosemeters. (authors)

  10. Development and underground testing of the α dosimeter: a solid state electronic personal radiation dosimeter for uranium miners

    International Nuclear Information System (INIS)

    Parkinson, R.N.; Roze, V.; Shepherd, R.

    1981-01-01

    The αDOSIMETER is a complete, integrated system designed to monitor the immediate worksite of underground miners where the disintegration for radon daughters is a risk to the health of mining personnel. The dosimeter weighing little more than one pound is worn by each miner throughout the entire shift and is powered by the miner's cap lamp battery. After this integration period, the unit is connected to a reading network whereupon the day's data is dumped, calculated and stored. Beginning in July 1980, prototype units were subjected to vigorous underground testing in uranium mines in Canada and the United States and in tin mines in Cornwall, UK. The testing results are summarized and proposals advanced for a typical mine monitoring system utilizing the αDOSIMETER

  11. Provision of dosimeters by official monitoring services for eye lens dose estimation; Die Bereitstellung von Dosimetern durch die amtlichen Messstellen zur Abschaetzung der Augenlinsendosis

    Energy Technology Data Exchange (ETDEWEB)

    Engelhardt, J.; Martini, E. [Landesanstalt fuer Personendosimetrie und Strahlenschutzausbildung Mecklenburg-Vorpommern, Berlin (Germany)

    2013-07-01

    Recent epidemiological studies are implying that the radio sensitivity of the eye lens is much higher than supposed in the past. International recommendations and standards demand to lower down the annual limit of the eye lens organ dose to 20 mSv. Since about 10 years German monitoring services offer partial-body dosimeters fixed on the head or on glasses for monitoring the eye lens dose. These dosimeters are optimized to measure the (surface) personal dose equivalent H{sub p}(0,07) from 0,5 mSv up to 10 Sv, which clearly overestimate the organ dose of the eye lens. With special features like different calibrations partial-body dosimeters should be applicable for legal dosimetry to avoid the development of special H{sub p}(3) dosimeters. Accepting the right way for wearing these dosimeters it is important to get the right results. Practical experiences are shown with measuring results and the difficulties of rounding the exact measuring values to discrete dose steps. Closing this article we point to still missing legal basis and open questions regarding to type testing procedures. (orig.)

  12. Copper doped borate dosimeters revisited

    Energy Technology Data Exchange (ETDEWEB)

    Alajerami, Y.S.M. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Department of Medical Radiography, Al-Azhar University, Gaza Strip, Palestine (Country Unknown); Hashim, S., E-mail: suhairul@utm.my [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Oncology Treatment Centre, Sultan Ismail Hospital, 81100 Johor Bahru (Malaysia); Ghoshal, S.K. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Bradley, D.A. [Centre for Nuclear and Radiation Physics, Department of Physics, University of Surrey, Guildford GU2 7XH (United Kingdom); Department of Physics, Faculty of Science, University of Malaya, 50603 Kuala Lumpur (Malaysia); Mhareb, M. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); Saleh, M.A. [Department of Physics, Universiti Teknologi Malaysia, 81310 Skudai, Johor (Malaysia); National Atomic Energy Commission (NATEC), Sana' a (Yemen)

    2014-11-15

    We render a panoramic overview on copper (Cu) doped borate dosimeters. Preparing a dosimeter by mixing specific materials with precise weights and methods is a never-ending quest. The recommended composition is highly decisive for accurate estimation of the absorbed dose, prediction of the biological outcome, determination of the treatment dose for radiation therapy and facilitation of personal monitoring. Based on these principles, the proposed dosimeter must cover a series of dosimetric properties to realize the exact results and assessment. The doped borate dosimeters indeed demonstrate attractive thermoluminescence (TL) features. Several dedicated efforts are attempted to improve the luminescence properties by doping various transition metals or rare-earth elements. The Cu ion being one of the preferred activators shows excellent TL properties as revealed via detail comparison with other dosimeters. Two oxide states of Cu (Cu{sup +} and Cu{sup ++}) with reasonable atomic number allow easy interaction with boron network. Interestingly, the intrinsic luminescent centers of borate lattice are in cross linked with that of Cu{sup +} ions. Thus, the activation of borate dosimeter with Cu ions for the enhancement of the TL sensitivity is recognized. These dosimeters reveal similar glow curves as the standard TLD-100 (LiF:Mg,Ti) one irrespective of the use of modifiers and synthesis techniques. They display high sensitivity, low fading, dose response linearity over wide range and practical minimum detectable dose. Furthermore, the effective atomic number being the most beneficial aspect (equivalent to that of human tissue) of borate dosimeters do not show any change due to Cu ion activations. The past development, major challenges, excitement, applications, recent progress and the future promises of Cu doped borate TL dosimeters are highlighted. - Highlights: • The manuscript gives a panoramic overview on copper doped borate dosimeters. • Cu ions activated

  13. Experimental determination of the PTW 60019 microDiamond dosimeter active area and volume

    Energy Technology Data Exchange (ETDEWEB)

    Marinelli, Marco, E-mail: marco.marinelli@uniroma2.it; Prestopino, G.; Verona, C.; Verona-Rinati, G. [INFN–Dipartimento di Ingegneria Industriale, Università di Roma “Tor Vergata,” Via del Politecnico 1, Roma 00133 (Italy)

    2016-09-15

    Purpose: Small field output correction factors have been studied by several research groups for the PTW 60019 microDiamond (MD) dosimeter, by comparing the response of such a device with both reference dosimeters and Monte Carlo simulations. A general good agreement is observed for field sizes down to about 1 cm. However, evident inconsistencies can be noticed when comparing some experimental results and Monte Carlo simulations obtained for smaller irradiation fields. This issue was tentatively attributed by some authors to unintentional large variations of the MD active surface area. The aim of the present study is a nondestructive experimental determination of the MD active surface area and active volume. Methods: Ten MD dosimeters, one MD prototype, and three synthetic diamond samples were investigated in the present work. 2D maps of the MD response were recorded under scanned soft x-ray microbeam irradiation, leading to an experimental determination of the device active surface area. Profiles of the device responses were measured as well. In order to evaluate the MD active volume, the thickness of the diamond sensing layer was independently evaluated by capacitance measurements and alpha particle detection experiments. The MD sensitivity, measured at the PTW calibration laboratory, was also used to calculate the device active volume thickness. Results: An average active surface area diameter of (2.19 ± 0.02) mm was evaluated by 2D maps and response profiles of all the MDs. Average active volume thicknesses of (1.01 ± 0.13) μm and (0.97 ± 0.14) μm were derived by capacitance and sensitivity measurements, respectively. The obtained results are well in agreement with the nominal values reported in the manufacturer dosimeter specifications. A homogeneous response was observed over the whole device active area. Besides the one from the device active volume, no contributions from other components of the housing nor from encapsulation materials were observed in

  14. Evaluation of personal integrating dosimeters

    International Nuclear Information System (INIS)

    Correa, C.A.; Bisauta, Mauricio A.

    2007-01-01

    The objective of this work is to analyze the different types of dosimeters present in the international market that are used to provide personal dose monitoring, specifically for external gamma and beta radiation, Hp(10) and Hp (0,07), as well as to add comments of advances in the field of passive and operative dosimetry, and the changes that are being produced in the regulating policy in other countries regarding the use of this devices. The technical specification of each dosimeter has been extracted of different catalogues of products. To conclude, the importance has been stressed in a proper selection of dosimeters with its advantages and disadvantages before its use. (author) [es

  15. Development of personal dose monitoring system using wireless data transmission device

    International Nuclear Information System (INIS)

    Inui, Daisuke; Nakashima, Sadao

    2008-01-01

    Radiation workers working in radiation controlled area in nuclear power plants etc., are required to carry a dosimeters by regulation law. The workers are controlled daily on personal exposure dose by reading out the exposure dose information of the dosimeters with an area access control gate installed at the entrance of the radiation controlled area. This type of personal dose monitoring system has a problem that each worker can get his personal dose data only at the entrance of the radiation controlled area several times a day. We developed a system to get the real-time acquisition of personal dose data especially for workers working in a high dose area. This system is generally composed of a dosimeter with a wireless attachment, relay station, and monitor. Some relay stations set in main work places in the radiation controlled area can collect real-time personal dose data of each dosimeter carried by workers at the work place with the relay stations, and transmit it to the monitor to get personal dose data of individual workers. A wireless communication system between dosimeters and relay stations is applied to collect efficiently all personal dose data in the work place. (author)

  16. Plastic dosimeter

    International Nuclear Information System (INIS)

    Nagai, Shiro; Matsuda, Kohji.

    1988-01-01

    The report outlines major features and applications of plastic dosimeters. Some plastic dosimeters, including the CTA and PVC types, detect the response of the plastic material itself to radiations while others, such as pigment-added plastic dosimeters, contain additives as radiation detecting material. Most of these dosimeters make use of color centers produced in the dosimeter by radiations. The PMMA dosimeter is widely used in the field of radiation sterilization of food, feed and medical apparatus. The blue cellophane dosimeter is easy to handle if calibrated appropriately. The rad-color dosimeter serves to determine whether products have been irradiated appropriately. The CTA dosimeter has better damp proofing properties than the blue cellophane type. The pigment-added plastic dosimeter consists of a resin such as nylon, CTA or PVC that contains a dye. Some other plastic dosimeters are also described briefly. Though having many advantages, these plastic dosimeter have disadvantages as well. Some of their major disadvantages, including fading as well as large dependence on dose, temperature, humidity and anviroment, are discussed. (Nogami, K.)

  17. Operation of Personal Electronic Dosimeters at NRCN

    International Nuclear Information System (INIS)

    Weinstein, M.; Abraham, A.; Tshuva, A.; German, U.

    2004-01-01

    In the recent years, electronic personal dosimeters (EPD's) are increasingly being used at NRCN, replacing the old direct reading dosimeters that are still widely used. The most significant advantage of the new dosimeters is the real time alarm in a radiation field exceeding a pre-determined threshold. The EPD dosimeters are more precise and can measure γ, β and x rays of a wide range of energies. In addition, the electronic dosimeters collects and stores the reading at a fixed pattern (every 10 seconds) and keeps the data until it is downloaded from the dosimeter. This feature gives the ability to build a personal time-dependent exposure report for each worker who carries this device and to analyze, identify and measure the exact dose, time and duration of any exposure event he was involved in. Designing and building a personal electronic dosimeter became possible as a result of the massive technological improvements of semi conductor detectors and the minimization processes of microprocessors and low energy electronic devices. The main purpose for personal electronic dosimeters was to monitor on-line doses for radiation workers.A special reader device enables to download data and upload operational settings of the dosimeters. By means of this communication channel, one can save the data acquired by the dosimeter, clear the dosimeter memory and set the dosimeter operational parameters. There are two possible working patterns. The first is to read and set all the dosimeters at a central point, normally a dosimetry laboratory (single reader) and the second and more expensive one, is to build a network of readers covering the plant for obtaining on-line communication

  18. Area radiation monitoring on ISS Increments 17 to 22 using PADLES in the Japanese Experiment Module Kibo

    International Nuclear Information System (INIS)

    Nagamatsu, A.; Murakami, K.; Kitajo, K.; Shimada, K.; Kumagai, H.; Tawara, H.

    2013-01-01

    The measurement of radiation environmental parameters in space is essential to support radiation risk assessments for astronauts and establish a benchmark for space radiation models for present and future human space activities. The Japan Aerospace Exploration Agency (JAXA) is performing a continuous area radiation monitoring experiment using the “PAssive Dosimeters for Lifescience Experiments in Space” (PADLES) system inside the Japanese Experiment Module Kibo on board the International Space Station (ISS). The PADLES dosimeter consists of thermoluminescent dosimeters (TLDs) and CR-39 plastic nuclear track detectors (PNTDs). JAXA has run the Area PADLES experiment since the Kibo module was attached to the ISS in June 2008, using 17 dosimeters in fixed locations on the Pressurized Module (PM) and the Experiment Logistics Module-Pressurized Section (ELM-PS) of Kibo, which are replaced every 6 months or every Increment, respectively. For three monitoring periods, known as Area PADLES experiment series no. 1 to no. 3, of 301, 180, and 232 days in June 2008 to April 2010 over ISS Increments 17 to 22, the average absorbed dose (dose equivalent) rates of 12 positions in the PM of Kibo were 319 ± 30 μGy/day (618 ± 102 μSv/day), 276 ± 30 μGy/day (608 ± 94 μSv/day), and 293 ± 33 μGy/day (588 ± 84 μSv/day), respectively. The radiation measurement in the ELM-PS was conducted in only Area PADLES experiment series no. 3 from August 2009 to April 2010 (232 days) over ISS Increments 21 to 22, the average absorbed dose (dose equivalent) rates of 5 positions was 297 ± 28 μGy/day (661 ± 65 μSv/day). The directional dependence of the radiation field was also investigated by installing PADLES dosimeters located in the zenith of ELM-PS of Kibo. -- Highlights: • This article shows the first results of dose measurement inside the Japanese Experiment Module Kibo with the PADLES system. • Generating spatial dose distribution data with the PADLES system are key

  19. Stable Chemical Dosimeters for Partial Reconstruction of Nuclear Accident Conditions

    Energy Technology Data Exchange (ETDEWEB)

    Dvornik, I.; Zec, U.; Baric, M.; Razem, D. [Ruder Boskovic Nuclear Institute, Zagreb, Yugoslavia (Croatia)

    1969-10-15

    The application of chemical dosimeters, tissue equivalent with respect to gamma rays and neutrons, is proposed for dosimetric topography of the space around nuclear devices in case of accidents. The dosimeters in the form of sealed glass ampoules have sufficient sensitivity and long-term stability and are evaluated or checked directly by conventional spectrophotometry. The sensitivity, expressed as yield per rad, is approximately equal for gamma rays and neutrons. The resolution in both cases is about one rad, and the range is up to several thousand rads. The precision of dosimetry is {+-} 1 rad or {+-} 2%, whichever is higher. In free space and unshielded the dosimeter measures the total rad-absorbed dose delivered by gamma rays and neutrons, i.e. the first collision gamma plus neutron dose. If used on- or in-phantom, especially if several dosimeters are disposed within and around the same phantom, it can give important data about the amount of the neutron component of the dose and about the effective mean energy of incident neutrons. The neutron component of the dose can be directly measured if the gamma dosimeter is used together with the chemical dosimeter. The experiments giving the change of optical density per rad and the radiation chemical yield with respect to the absorbed dose delivered by 14-MeV neutrons are described in detail. The possibility is also mentioned of applying the dosimeter as a very sensitive monitor for thermal neutrons, which is due to the chlorine content of 4.73% and activation to {sup 38}Cl. The opinion is expressed that this dosimeter deserves some attention as a part of future planning and development work on area and personnel accidental dosimetry systems. (author)

  20. Characterization of a Cs-137 radiation beam for dosimeter calibrations in the CRCN-CO

    International Nuclear Information System (INIS)

    Baptista Neto, Annibal T.; Soares, Carlos M. de A.; Silva, Teogenes A. da; Correa, Rosangela da S.

    2009-01-01

    The Centro Regional de Ciencias Nucleares do Centro Oeste (CRCN-CO) has played an important role in the environmental radiation monitoring program in Goiania city. The reduce its dependence of others monitoring laboratories, the CRCN-CO acquired a model 28 JL Shepherd and Associates irradiation system with a 137 Cs source for calibrations and frequent quality control checks of radiation dosimeters. A characterization of the irradiation system was carried out with the reference standard dosimeters that are maintained by the Centro de Desenvolvimento da Tecnologia Nuclear (CDTN). The laboratory surrounding areas were monitored to demonstrate the adequate radiation protection conditions and parameters as the radiation field size, beam uniformity and the level of the scatter radiation were investigated. Dosimetry of the 137 Cs radiation beam in terms of air kerma rate was carried out at many source-detector distances with 4 (four) different beam lead attenuators. Results demonstrated that in spite of the radiation shutter automation is strongly recommended, the irradiation system is adequate and it complies with the requirements to be used for dosimeter irradiations and calibrations for the purpose of radiation protection. (author)

  1. A radon (thoron) daughter personal alpha-dosimeter of the passive type using a diffused-junction detector and an electrostatic collector

    International Nuclear Information System (INIS)

    Bigu, J.; Frattini, A.

    1984-05-01

    A solid-state alpha-dosimeter has been designed and found to be suitable for personal and environmental radon-thoron daughter monitoring. The dosimeter basically consists of an electrostatic collector and an alpha-particle counting system with spectroscopy capabilities. The sensitive volume (∼20 cm 3 ) of the electrostatic collector consists of a cylindrically-shaped metal wire screen and a diffused-junction silicon alpha-detector covered with a thin aluminized mylar sheet. A DC voltage (∼450 V) is applied between the wire screen and the mylar sheet, the latter held at negative potential relative to the metal screen. Data can be retrieved during or after sampling by means of a micro-computer (Epson HX20) via a RS-232 communcation interface unit. The dosimeter has been calibrated in a large (26 m 3 ) radon/thoron test facility. A linear relationship was found between radon gas concentration and radon daughter Working Level, and the dosimeter's alpha-count. The dosimeter is mounted on top of an ordinary miner's cap lamp battery and is ideally suited for personal monitoring in underground uranium mines and other working areas. The dosimeter presented here is a considerably improved version of an earlier prototype

  2. Development of neutron dosimeter using CR-39 for measurement of ambient dose equivalent

    International Nuclear Information System (INIS)

    Maki, Daisuke; Shinozaki, Wakako; Ohguchi, Hiroyuki; Yamamoto, Takayoshi; Nakamura, Takayoshi

    2010-01-01

    A CR-39 has good advantages such as cumulative type dosimeter, small fading effect and gamma-ray insensitive. Therefore, we developed the wide energy-range environmental neutron dosimeter using eight CR-39s for area monitoring in this study. This dosimeter is made of octagonal columnar polyethylene block which height is 60 mm and bottom side is 25 mm. The dosimeter contains two types of CR-39s for fast neutron detection and slow neutron detection. Four CR-39s for fast neutron detection are used for detection of recoil protons produced by H (n, p) reactions. Four CR-39s for slow neutron detection are used with boron nitride converter to detect alpha-rays produced by 10 B (n, α) 7 Li reactions. Ambient dose equivalent is obtained by adding the number of etch-pits observed in four CR-39s for fast neutron detection to the number of etch-pits observed in four CR-39s for slow neutron detection with appropriate constants respectively. Dosimeters were irradiated with some energetic neutrons and evaluated results of ambient dose equivalent were compared with results from neutron transport calculations. Energy response of dosimeter shows good agreement with neutron fluence to ambient dose equivalent conversion coefficients. Directional dependence of dosimeter is at the same level as the rem-counter. (author)

  3. Citizen's dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Klemic, Gladys [Naperville, IL; Bailey, Paul [Chicago, IL; Breheny, Cecilia [Yonkers, NY

    2008-09-02

    The present invention relates to a citizen's dosimeter. More specifically, the invention relates to a small, portable, personal dosimetry device designed to be used in the wake of a event involving a Radiological Dispersal Device (RDD), Improvised Nuclear Device (IND), or other event resulting in the contamination of large area with radioactive material or where on site personal dosimetry is required. The card sized dosimeter generally comprises: a lower card layer, the lower card body having an inner and outer side; a upper card layer, the layer card having an inner and outer side; an optically stimulated luminescent material (OSLM), wherein the OSLM is sandwiched between the inner side of the lower card layer and the inner side of the upper card layer during dosimeter radiation recording, a shutter means for exposing at least one side of the OSLM for dosimeter readout; and an energy compensation filter attached to the outer sides of the lower and upper card layers.

  4. Calibration and performance testing of electronic personal dosimeters (EPD)

    International Nuclear Information System (INIS)

    Banaga, H.A.

    2008-04-01

    In modern radiation protection practices, active personal dosimeters are becoming absolutely necessary operational tools for satisfying the ALARA principle. The aim of this work was to carry out calibration and performance testing of ten electronic personal dosimeters (EPD) used for the individual monitoring. The EPDs were calibrated in terms of operation radiation protection quantity, personal dose equivalent, Hp (10). Calibrations were carried out at three of x-ray beam qualities described in ISO 4037 namely 60, 100 and 150 kV in addition to Cs-137 gamma ray quality. The calibrations were performed using polymethylmethacrylate (PMMA) phantom with dimensions 20*20*15 cm 3 . Conversion coefficient Hp (10)/K air for the phantom was also calculated. The response and linearity of the dosimeter at the specified energies were also tested. The EPDs tested showed that the calibration coefficient ranged from 0.60 to 1.31 and an equivalent response for the specified energies that ranged from 0.76 to 1.67. The study demonstrated the possibility of using non standard phantom for calibrating dosimeters used for individual monitoring. The dosimeters under study showed a good response in all energies except the response in quality 100 kV. The linearity of the dosimeters was within ±15%, with the exception of the quality 100 kV where this limit was exceeded.(Author)

  5. Radon daughter dosimeter

    International Nuclear Information System (INIS)

    Durkin, J.

    1977-01-01

    This patent describes a portable radon daughter dosimeter unit used to measure radon gas alpha daughters in ambient air. These measurements can then be related to preselected preestablished standards contained in a remote central readout unit. The dosimeter unit is adapted to be worn by an operator in areas having alpha particle radiation such as in uranium mines. Within the dosimeter is a detector head housing having a filter head and a solid state surface barrier radiation detector; an air pump to get air to the detector head; a self contained portable power supply for the unit; and electronic circuitry to process detected charged electrons from the detector head to convert and count their pulses representatives of two alpha radon emitter daughters. These counted pulses are in binary form and are sent to a readout unit where a numerical readout displays the result in terms of working level-hours

  6. Radon daughter dosimeter

    International Nuclear Information System (INIS)

    Durkin, J.

    1977-01-01

    A portable radon daughter dosimeter unit used to measure Radon gas alpha daughters in ambient air is described. These measurements can then be related to preselected preestablished standards contained in a remote central readout unit. The dosimeter unit is adapted to be worn by an operator in areas having alpha particle radiation such as uranium mines. Within the dosimeter is a detector head housing having a filter head and a solid state surface barrier radiation detector; an air pump to get air to the detector head; a self contained portable power supply for the unit; and electronic circuitry to process detected charged electrons from the detector head to convert and count their pulses representatives of two alpha radon emitter daughters. These counted pulses are in binary form and are sent to a readout unit where a numerical readout diplays the result in terms of working level-hours

  7. Chemical dosimeter

    International Nuclear Information System (INIS)

    Baker, W.B.; Clark, D.G.

    1979-01-01

    The dosimeter may be carried by individuals e.g. at the belt and serves to monitor for vinyl-chloride vapors in industrial plants and for toxic radon gas and toxic radon gas products in mines. It contains a pump, sucking an air flow through an orifice and a filter, as well as a sensor circuit for detecting low air flow rates and a battery testing circuit. (DG) 891 HP/DG 892 MKO [de

  8. Personnel monitoring

    International Nuclear Information System (INIS)

    1965-01-01

    This film stresses the need for personnel monitoring in work areas where there is a hazard of exposure to radiation. It illustrates the use of personnel monitoring devices (specially the film dosimeter), the assessment of exposure to radiation and the detailed recording of the results on personnel filing cards

  9. Personnel monitoring

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    1966-12-31

    This film stresses the need for personnel monitoring in work areas where there is a hazard of exposure to radiation. It illustrates the use of personnel monitoring devices (specially the film dosimeter), the assessment of exposure to radiation and the detailed recording of the results on personnel filing cards

  10. SSDL Preparation for Implementation of the Use of OSL Dosimeters in Malaysia

    International Nuclear Information System (INIS)

    Sangau, J.K.; Taiman Kadni; Ahmad Bazlie Abdul Kadir

    2013-01-01

    Since the early 1980's, film badge has been widely used as a device of personal dose monitoring in Malaysia. Secondary Standard Dosimetry Laboratory (SSDL), as a service center for film badge has obtained the supply of personal monitoring film from Agfa Gevaert, Belgium every year. As the uses of film badge have some weaknesses, it has prompted SSDL to find an alternative dosimeter to replace the film badge. Based on the studies that have been conducted, SSDL has selected OSL dosimeter (Optically Stimulated Luminescent Dosimeter) to replace the film badge and is expected to be fully operational by middle of 2015. This paper aims to explain the selection of OSL dosimeter and planning carried out to ensure the success of their application in Malaysia. (author)

  11. The dosimeter personal use in controlled area

    International Nuclear Information System (INIS)

    Costa, R. F.

    2015-01-01

    The discovery of X-rays revolutionized medicine because it allowed a patient to be examined internally with no surgery. But also caused damage to health professionals and patients due, its oxidizing action. In the beginning of its discovery, many doctors were exposed and exposed beams to their patients for long periods of time, therefore, they developed diseases caused by radiation and the medical community realized that something was wrong. Then created a radiological protection commission to regulate its use in humans and so limit your exposure. Today we know that many companies still did not fit the standards of radiation protection. So we evaluate the technical professionals in radiology regarding the correct use of personal dosimeter, through a descriptive study with a quantitative approach, we used the information collection technique based on a questionnaire developed for this purpose which was delivered and collected personally. From this survey, we sought to assess the knowledge of the basic guidelines of radiological protection. He concluded that the majority of respondents know the rules of use of the personal dosimeter, but do not use it properly, due mainly to lack of supervision by the company, overwork and neglect. (author)

  12. Composite material dosimeters

    Science.gov (United States)

    Miller, Steven D.

    1996-01-01

    The present invention is a composite material containing a mix of dosimeter material powder and a polymer powder wherein the polymer is transparent to the photon emission of the dosimeter material powder. By mixing dosimeter material powder with polymer powder, less dosimeter material is needed compared to a monolithic dosimeter material chip. Interrogation is done with excitation by visible light.

  13. Floating Gate CMOS Dosimeter With Frequency Output

    Science.gov (United States)

    Garcia-Moreno, E.; Isern, E.; Roca, M.; Picos, R.; Font, J.; Cesari, J.; Pineda, A.

    2012-04-01

    This paper presents a gamma radiation dosimeter based on a floating gate sensor. The sensor is coupled with a signal processing circuitry, which furnishes a square wave output signal, the frequency of which depends on the total dose. Like any other floating gate dosimeter, it exhibits zero bias operation and reprogramming capabilities. The dosimeter has been designed in a standard 0.6 m CMOS technology. The whole dosimeter occupies a silicon area of 450 m250 m. The initial sensitivity to a radiation dose is Hz/rad, and to temperature and supply voltage is kHz/°C and 0.067 kHz/mV, respectively. The lowest detectable dose is less than 1 rad.

  14. Neutron area monitoring at storage bunkers of density/moisture gauges

    Energy Technology Data Exchange (ETDEWEB)

    Garcia-Fuste, M.J. [Grup de Fisica de les Radiacions. Universitat Autonoma de Barcelona (UAB), Edifici Cc, E-08193 Bellaterra (Spain); Amgarou, K., E-mail: khalil.amgarou@uab.ca [Grup de Fisica de les Radiacions. Universitat Autonoma de Barcelona (UAB), Edifici Cc, E-08193 Bellaterra (Spain); Garcia-Orellana, J.; Domingo, C. [Grup de Fisica de les Radiacions. Universitat Autonoma de Barcelona (UAB), Edifici Cc, E-08193 Bellaterra (Spain)

    2010-12-15

    Previous studies remarked the need of performing neutron personal dosimetry, together with gamma dosimetry, when using portable density/moisture gauges that are normally equipped with gamma and neutron sources. The convenience of our Poly-Allyl-Diglycol-Carbonate (PADC) based neutron dosimeter to perform long-term routine survey of all workers involved in the transport, maintenance and usage of these gauges was also corroborated in the past. This complete dosimetric control offers the possibility to quantify simultaneously individual neutron and gamma doses of workers operating such devices, especially in the most frequent cleaning and maintenance (shutter lubrication) operations of the gauges or in emergency situations that may lead to radiological risk of the persons involved. Another aspect to be considered, from the radioprotection point of view, is the optimization of the occupational and public exposure in the storage bunkers of the gauges. In this work, three storage bunkers, located at three different factories of the PAYMA Cotas S.A.U company, have been monitored for three months using several units of our PADC based neutron dosimeter. Special care has been taken to study also offices near the bunkers with high occupancy rates and passage zones, such as toilets and access areas. Results for all monitored points inside and around the three storage bunkers are presented. Neutron doses inside the bunkers could be relatively high depending on the specific conditions (geometry configuration, localization and shielding composition) of the considered bunker and the number of portable gauges stored.

  15. Neutron area monitoring at storage bunkers of density/moisture gauges

    International Nuclear Information System (INIS)

    Garcia-Fuste, M.J.; Amgarou, K.; Garcia-Orellana, J.; Domingo, C.

    2010-01-01

    Previous studies remarked the need of performing neutron personal dosimetry, together with gamma dosimetry, when using portable density/moisture gauges that are normally equipped with gamma and neutron sources. The convenience of our Poly-Allyl-Diglycol-Carbonate (PADC) based neutron dosimeter to perform long-term routine survey of all workers involved in the transport, maintenance and usage of these gauges was also corroborated in the past. This complete dosimetric control offers the possibility to quantify simultaneously individual neutron and gamma doses of workers operating such devices, especially in the most frequent cleaning and maintenance (shutter lubrication) operations of the gauges or in emergency situations that may lead to radiological risk of the persons involved. Another aspect to be considered, from the radioprotection point of view, is the optimization of the occupational and public exposure in the storage bunkers of the gauges. In this work, three storage bunkers, located at three different factories of the PAYMA Cotas S.A.U company, have been monitored for three months using several units of our PADC based neutron dosimeter. Special care has been taken to study also offices near the bunkers with high occupancy rates and passage zones, such as toilets and access areas. Results for all monitored points inside and around the three storage bunkers are presented. Neutron doses inside the bunkers could be relatively high depending on the specific conditions (geometry configuration, localization and shielding composition) of the considered bunker and the number of portable gauges stored.

  16. Radiation dosimeter

    International Nuclear Information System (INIS)

    Lowe, D.

    1980-01-01

    A radiation dosimeter is described, comprising a thermoluminescent phosphor incorporated in matrix of polyethersulphone. The dosimeter is preferably a thin film formed by spreading a suspension of a powdered phosphor in a solution of polyethersulphone onto a flat surface. The solvent for the polyethersulphone is a mixture of a n-methyl-2-pyrrolidone and xylene in equal proportions. A thin, inert film of polyethersulphone can be cemented to one surface of the dosimeter so as to provide a skin dosimeter. (author)

  17. Performance and type testing of selected dosimeters used for individual monitoring

    International Nuclear Information System (INIS)

    Almhena, E. H. Y.

    2010-07-01

    This study describes calibration and performance testing carried out for a set of 14 electronic personal dosimeters (EPDs) and thermoluminescence dosimeters (TLDs ) at the Secondary Standard Dosimetry Laboratory of Sudan. Also the conversion coefficients from air kerma have been determined. Dosimeters tested are belonging to three manufactures representing most commonly used types in Sudan. Calibrations were carried out at three X-ray qualities: 80, 120 and 150 kV, ISO 'narrow' spectra and for ''1''3''7Cs, '' 60 Co gamma rays. The experiments were carried out using ICRU Slab phantom with dimension 30x30x15 cm. Secondary standard ionization chamber was used to measure the personal dose equivalent, Hp(10) as standard dosimetric quantity of interest. parameters tested include: The variation of response with radiation energy and angle of incident in addition to dose rate dependence. The angular dependence factors have been experimentally determined for the same qualities and for different angles (0, ±20, ±40, ±60u) + were performed in accordance to the relevant international standards. Excellent energy, angular and dose rate response was demonstrated for 662 137 Cs, 1250 60 Co beam and (118, 100,65 keV) x-ray beam qualities exception the EPD at PM1203M are slightly larger but still with the acceptable. The response of electronic dosimeters were found to in limits of acceptable performance. For the mean response of all energies is range of EPDs Type RADOS 60, Greatz ED 150, Polimaster PM1203M, TLD were (0.75 ±0.08- 1.13±0.13), (0.83±0.29 -1.06±0.07), (1.08±0.14-1.27±0.09), (0.99±0.05 - 1.14±0.13) respectively. The majority of the dosimeters tested showed good energy and angular response. (Author)

  18. Dosimeter charging apparatus

    International Nuclear Information System (INIS)

    Reuter, F.A.; Moorman, Ch.J.

    1985-01-01

    An apparatus for charging a dosimeter which has a capacitor connected between first and second electrodes and a movable electrode in a chamber electrically connected to the first electrode. The movable electrode deflects varying amounts depending upon the charge present on said capacitor. The charger apparatus includes first and second charger electrodes couplable to the first and second dosimeter electrodes. To charge the dosimeter, it is urged downwardly into a charging socket on the charger apparatus. The second dosimeter electrode, which is the dosimeter housing, is electrically coupled to the second charger electrode through a conductive ring which is urged upwardly by a spring. As the dosimeter is urged into the socket, the ring moves downwardly, in contact with the second charger electrode. As the dosimeter is further urged downwardly, the first dosimeter electrode and first charger electrode contact one another, and an insulator post carrying the first and second charger electrodes is urged downwardly. Downward movement of the post effects the application of a charging potential between the first and second charger electrodes. After the charging potential has been applied, the dosimeter is moved further into the charging socket against the force of a relatively heavy biasing spring until the dosimeter reaches a mechanical stop in the charging socket

  19. Offsite environmental monitoring report. Radiation monitoring around United States nuclear test areas, calendar year 1997

    International Nuclear Information System (INIS)

    Davis, M.G.; Flotard, R.D.; Fontana, C.A.; Hennessey, P.A.; Maunu, H.K.; Mouck, T.L.; Mullen, A.A.; Sells, M.D.

    1999-01-01

    This report describes the Offsite Radiological Environmental Monitoring Program (OREMP) conducted during 1997 by the US Environmental Protection Agency's (EPAs), Radiation and Indoor Environments National Laboratory, Las Vegas, Nevada. This laboratory operated an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling and analyzing milk, water, and air; by deploying and reading thermoluminescent dosimeters (TLDs); and using pressurized ionization chambers (PICs) to measure ambient gamma exposure rates with a sensitivity capable of detecting low level exposures not detected by other monitoring methods

  20. Offsite environmental monitoring report: Radiation monitoring around United States nuclear test areas, calendar year 1997

    Energy Technology Data Exchange (ETDEWEB)

    Davis, M.G.; Flotard, R.D.; Fontana, C.A.; Hennessey, P.A.; Maunu, H.K.; Mouck, T.L.; Mullen, A.A.; Sells, M.D.

    1999-01-01

    This report describes the Offsite Radiological Environmental Monitoring Program (OREMP) conducted during 1997 by the US Environmental Protection Agency`s (EPAs), Radiation and Indoor Environments National Laboratory, Las Vegas, Nevada. This laboratory operated an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling and analyzing milk, water, and air; by deploying and reading thermoluminescent dosimeters (TLDs); and using pressurized ionization chambers (PICs) to measure ambient gamma exposure rates with a sensitivity capable of detecting low level exposures not detected by other monitoring methods.

  1. Offsite environmental monitoring report: Radiation monitoring around United States nuclear test areas, calendar year 1993

    International Nuclear Information System (INIS)

    Chaloud, D.J; Daigler, D.M.; Davis, M.G.

    1996-06-01

    This report describes the Offsite Radiation Safety Program conducted during 1993 by the Environmental Protection Agency's (EPA's) Environmental Monitoring Systems Laboratory - Las Vegas (EMSL-LV). This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ionization chambers (PICs); by biological monitoring of foodstuffs including animal tissues and food crops; and by measurement of radioactive material deposited in humans

  2. Offsite environmental monitoring report: Radiation monitoring around United States nuclear test areas, calendar year 1993

    Energy Technology Data Exchange (ETDEWEB)

    Chaloud, D.J; Daigler, D.M.; Davis, M.G. [and others

    1996-06-01

    This report describes the Offsite Radiation Safety Program conducted during 1993 by the Environmental Protection Agency`s (EPA`s) Environmental Monitoring Systems Laboratory - Las Vegas (EMSL-LV). This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ionization chambers (PICs); by biological monitoring of foodstuffs including animal tissues and food crops; and by measurement of radioactive material deposited in humans.

  3. Electron beam energy monitoring using thermoluminescent dosimeters and electron back scattering

    International Nuclear Information System (INIS)

    Nelson, Vinod; Gray, Alison

    2013-01-01

    Periodic checks of megavoltage electron beam quality are a fundamental requirement in ensuring accurate radiotherapy treatment delivery. In the present work, thermoluminescent dosimeters (TLDs) positioned on either side of a lead sheet at the surface of a water equivalent phantom were used to monitor electron beam quality using the electron backscattering method. TLD100 and TLD100H were evaluated as upstream detectors and TLD200, TLD400 and TLD500 were evaluated as downstream detectors. The evaluation assessed the test sensitivity and correlation, long and short term reproducibility, dose dependence and glow curve features. A prototype of an in-air jig suitable for use in postal TLD dose audits was also developed and an initial evaluation performed. The results indicate that the TLD100-TLD200 combination provides a sensitive and reproducible method to monitor electron beam quality. The light weight and easily fabricated in-air jig was found to produce acceptable results and has the potential to be used by radiation monitoring agencies to carry out TLD postal quality assurance audits, similar to audits presently being conducted for photon beams. -- Highlights: ► Monitoring electron beam quality via electron backscattering was investigated. ► Different thermoluminescent materials were evaluated as detectors. ► A TLD100-TLD200 combination produced the most sensitive and reproducible results. ► An in-air jig was evaluated to allow measurements via postal dose audits

  4. The development of remote wireless radiation dose monitoring system

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Jin-woo [KAERI - Korea Atomic Energy Research Institute, Jeongup-si (Korea, Republic of); Chonbuk National University, Jeonjoo-Si (Korea, Republic of); Jeong, Kyu-hwan [KINS - Korea Institute of Nuclear Safety, Daejeon-Si (Korea, Republic of); Kim, Jong-il [Chonbuk National University, Jeonjoo-Si (Korea, Republic of); Im, Chae-wan [REMTECH, Seoul-Si (Korea, Republic of)

    2015-07-01

    Internet of things (IoT) technology has recently shown a large flow of IT trends in human life. In particular, our lives are now becoming integrated with a lot of items around the 'smart-phone' with IoT, including Bluetooth, Near Field Communication (NFC), Beacons, WiFi, and Global Positioning System (GPS). Our project focuses on the interconnection of radiation dosimetry and IoT technology. The radiation workers at a nuclear facility should hold personal dosimeters such as a Thermo-Luminescence Dosimeter (TLD), an Optically Stimulated Luminescence Dosimeter (OSL), pocket ionization chamber dosimeters, an Electronic Personal Dosimeter (EPD), or an alarm dosimeter on their body. Some of them have functions that generate audible or visible alarms to radiation workers in a real working area. However, such devices used in radiation fields these days have no functions for communicating with other areas or the responsible personnel in real time. In particular, when conducting a particular task in a high dose area, or a number of repair works within a radiation field, radiation dose monitoring is important for the health of the workers and the work efficiency. Our project aims at the development of a remote wireless radiation dose monitoring system (RWRD) that can be used to monitor the radiation dose in a nuclear facility for radiation workers and a radiation protection program In this project, a radiation dosimeter is the detection device for personal radiation dose, a smart phone is the mobile wireless communication tool, and, Beacon is the wireless starter for the detection, communication, and position of the worker using BLE (Bluetooth Low Energy). In this report, we report the design of the RWRD and a demonstration case in a real radiation field. (authors)

  5. The development of remote wireless radiation dose monitoring system

    International Nuclear Information System (INIS)

    Lee, Jin-woo; Jeong, Kyu-hwan; Kim, Jong-il; Im, Chae-wan

    2015-01-01

    Internet of things (IoT) technology has recently shown a large flow of IT trends in human life. In particular, our lives are now becoming integrated with a lot of items around the 'smart-phone' with IoT, including Bluetooth, Near Field Communication (NFC), Beacons, WiFi, and Global Positioning System (GPS). Our project focuses on the interconnection of radiation dosimetry and IoT technology. The radiation workers at a nuclear facility should hold personal dosimeters such as a Thermo-Luminescence Dosimeter (TLD), an Optically Stimulated Luminescence Dosimeter (OSL), pocket ionization chamber dosimeters, an Electronic Personal Dosimeter (EPD), or an alarm dosimeter on their body. Some of them have functions that generate audible or visible alarms to radiation workers in a real working area. However, such devices used in radiation fields these days have no functions for communicating with other areas or the responsible personnel in real time. In particular, when conducting a particular task in a high dose area, or a number of repair works within a radiation field, radiation dose monitoring is important for the health of the workers and the work efficiency. Our project aims at the development of a remote wireless radiation dose monitoring system (RWRD) that can be used to monitor the radiation dose in a nuclear facility for radiation workers and a radiation protection program In this project, a radiation dosimeter is the detection device for personal radiation dose, a smart phone is the mobile wireless communication tool, and, Beacon is the wireless starter for the detection, communication, and position of the worker using BLE (Bluetooth Low Energy). In this report, we report the design of the RWRD and a demonstration case in a real radiation field. (authors)

  6. Wireless monitoring system for personal dose

    International Nuclear Information System (INIS)

    Kobayashi, Hironobu; Kawamura, Takeshi; Inoue, Takayuki

    2000-01-01

    Fuji Electric has developed a system for the higher radiation controlled area in nuclear power plants, in which exposure dose data measured on the wearer's chest, hands, and legs are transferred by wireless to the data control equipment so that the exposure dose can be controlled in real time. The system using a specified low-power radio wave causes no interference to the other types of dosimeters. The data control equipment automatically saves data received from the dosimeters and also has functions of calibration of dosimeters and maintenance of the wireless system. This paper describes the wireless monitoring system that consists of chest and parts dosimeters and data control equipment. (author)

  7. An approved personal dosimetry service based on an electronic dosimeter

    International Nuclear Information System (INIS)

    Marshall, T.O.; Bartlett, D.T.; Burgess, P.H.; Campbell, J.I.; Hill, C.E.; Pook, E.A.; Sandford, D.J.

    1991-01-01

    At the Second Conference on Radiation Protection and Dosimetry a paper was presented which, in part, announced the development of an electronic dosimeter to be undertaken in the UK by the National Radiological Protection Board (NRPB) and Siemens Plessey Controls Ltd. This dosimeter was to be of a standard suitable for use as the basis of an approved personal dosimetry service for photon and beta radiations. The project has progressed extremely well and dosimeters and readers are about to become commercially available. The system and the specification of the dosimeter are presented. The NRPB is in the process of applying for approval by the Health and Safety Executive (HSE) to operate as personal monitoring service based on this dosimeter. As part of the approval procedure the dosimeter is being type tested and is also undergoing an HSE performance test and wearer trials. The tests and the wearer trials are described and a summary of the results to date presented. The way in which the service will be organized and operated is described and a comparison is made between the running of the service and others based on passive dosimeters at NRPB

  8. Environmental gamma radiation monitoring at Visakhapatnam using thermoluminescence dosimeters

    International Nuclear Information System (INIS)

    Swarnkar, M.; Sahu, S.K.; Takale, R.A.; Shetty, P.G.; Pundit, G.G.; Puranik, V.D.

    2012-01-01

    The gamma rays are the most significant part of environmental dose due to its large range and deep penetrating power. The environmental gamma radiation is mainly originated from two sources natural radiation and artificially produced radiation. The natural radiation dose arises from the cosmic radiation (galactic and solar) and from the Earth (terrestrial) surface. In the last few decades there is a growing concern all over the world about radiation and their exposure to population. Thus it is necessary to conduct radiological environmental surveillance. The radiation survey data are useful to establish the natural background gamma radiation levels. Extensive gamma radiation survey was carried out around the surroundings of Vishakhapatnam using Thermoluminescence Dosimeters (TLDs). The CaSO 4 :(0.2 mole %) Dy Teflon TLD discs, specifically designed for environmental gamma radiation monitoring purpose were used. These TLD badge are having very high TL sensitivity, a negligible fading rate and a stable TL response. TLDs were deployed on quarterly basis for two years to obtain the cumulative gamma background radiation levels in the study area. The radiological survey was also carried out by using a calibrated radiation survey meter. The annual dose rates were computed from quarterly values actually found and normalised to 365 days. The environmental gamma radiation levels around Vishakhapatnam were found to be in the range of 0.79 mGy/y to 1.86 mGy/y. It is clearly seen from the results that location to location there is a large variation in external gamma radiation levels. During the cycle of the TLD survey, spot readings of the background radiation levels were taken, both while placing the TLDs and while removing them. The instantaneous dose rates measured using survey meter, are also following the large variation as found in TLDs. It varies between 110 nGy/hr to 210 nGy/hr. (author)

  9. Offsite environmental monitoring report: Radiation monitoring around United States nuclear test areas, calendar year 1991

    International Nuclear Information System (INIS)

    Chaloud, D.J.; Dicey, B.B.; Mullen, A.A.; Neale, A.C.; Sparks, A.R.; Fontana, C.A.; Carroll, L.D.; Phillips, W.G.; Smith, D.D.; Thome, D.J.

    1992-01-01

    This report describes the Offsite Radiation Safety Program conducted during 1991 by the Environmental Protection Agency's (EPA's) Environmental Monitoring Systems Laboratory-Las Vegas. This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ion chambers (PICs); and by biological monitoring of animals, food crops, and humans. Personnel with mobile monitoring equipment are placed in areas downwind from the test site prior to each nuclear weapons test to implement protective actions, provide immediate radiation monitoring, and obtain environmental samples rapidly after any occurrence of radioactivity release. Comparison of the measurements and sample analysis results with background levels and with appropriate standards and regulations indicated that there was no radioactivity detected offsite by the various EPA monitoring networks and no exposure above natural background to the population living in the vicinity of the NTS that could be attributed to current NTS activities. Annual and long-term trends were evaluated in the Noble Gas, Tritium, Milk Surveillance, Biomonitoring, TLD, PIC networks, and the Long-Term Hydrological Monitoring Program

  10. US progress on the development of CR-39 based neutron dosimeters

    International Nuclear Information System (INIS)

    Hadlock, D.E.

    1987-06-01

    Historically at US nuclear facilities, two types of personnel neutron dosimeters have been in routine use: nuclear track emulsion-Type A (NTA) film and thermoluminescent dosimeter (TLD)-albedo. Both of these dosimeters have energy-dependent responses. Therefore, the neutron energy spectra must be known, to interpret the dosimeter results properly. A new state-of-the-art dosimetry system has been developed within the US Department of Energy (US DOE) Personnel Neutron Dosimeter Evaluation and Upgrade Program. This system is called the combination thermoluminescent dosimeter/track etch dosimeter (TLD/TED). This paper briefly describes US DOE research currently being conducted to further enhance the TED portion of the combination TLD/TED system. The research areas involved include dose sensitivity, neutron energy range, specialized radiators, self-developing dosimeters, and neutron spectrometry. 1 fig., 1 tab

  11. Personnel dose equivalent monitoring at SLAC using lithium-fluoride TLD's [thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Jenkins, T.M.; Busick, D.D.

    1987-03-01

    TLD's replaced film badges in the early 1970's for all dose equivalent monitoring, both neutron and photon, and for all locations at SLAC. The photon TLD's, composed of Li-7 loaded teflon discs, are calibrated using conventional gamma-ray sources; i.e., Co-60, Cs-137, etc. For these TLD's a nominal value of 1 nC/mrem is used, and is independent of source energy for 100 keV to 3 MeV. Since measured dose equivalents at SLAC are only a small fraction of the allowable levels, it was not deemed necessary to develop neutron dosimeters which would measure dose equivalent accurately for all possible neutron spectra. Today, wallet TLD's, composed of pairs of Li-7 and Li-6 discs, are used, with the Li-6 measuring only thermal neutrons; i.e., they aren't moderated in any way to make them sensitive to neutrons with energies greater than thermal. The assumption is made that there is a correlation between thermal neutron fluences and fast neutron fluences around the research area where almost all neutron doses (exclusive of sealed sources) are received. The calibration factor for these Li-6 TLD's is 1 nC/mrem of fast neutrons. The method of determining the validity of this calibration is the subject of this note. 4 refs., 9 figs., 1 tab

  12. Development of colored alumilite dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Obara, Kenjiro; Shibanuma, Kiyoshi [Japan Atomic Energy Research Inst., Naka, Ibaraki (Japan). Naka Fusion Research Establishment; Yagi, Toshiaki [Japan Atomic Energy Research Inst., Takasaki, Gunma (Japan). Takasaki Radiation Chemistry Research Establishment; Yokoo, Noriko [Radiation Application Development Association, Tokai, Ibaraki (Japan)

    2003-03-01

    In the ITER (International Thermonuclear Experimental Reactor), in-vessel components such as blanket and divertor, which are installed in the vacuum vessel of the ITER, are maintained by remote handling equipment (RH equipment). The RH equipment for maintenance is operated under sever environmental conditions, such as high temperature (50{approx}100 degC), high gamma-ray radiation ({approx}1 kGy/h) in an atmosphere of inert gas or vacuum; therefore many components of the RH equipment must have a suitable radiation resistance efficiency for long time operation (10{approx}100 MGy). Typical components of the RH equipment have been extensively tested under an intensive gamma-ray radiation. Monitoring of the radiation dose of the components of the RH equipment is essential to control the operation period of the RH equipment considering radiation resistance. However, the maximum measurable radiation dose of the conventional dosimeters, such as ionization chamber, liquid, glass and plastic dosimeters are limited to be approximately 1MGy which is too low to monitor the RH equipment for the ITER. In addition, these conventional dosimeters do not involve sufficient radiation resistance against the high gamma-ray radiation as well as are not easy handling and low cost. Based on the above backgrounds, a new dosimeter with bleaching of an azo group dye to be applied to a radiation monitor has been developed for high gamma-ray radiation use. The colored alumilite dosimeter is composed of the azo group dye (-N=N-) in an anodic oxidation layer of aluminum alloy (Al{sub 2}O{sub 3}). It can monitor the radiation dose by measuring the change of the bleaching of azo dye in the Al{sub 2}O{sub 3} layer due to gamma-ray irradiation. The degree of bleaching is measured as the change of hue (color) and brightness based on the Munsell's colors with a three dimensional universe using spectrophotometer. In the tests, the dependencies such as colors, anodized layer thickness, type of azo

  13. Personnel ionizing radiation dosimeter

    International Nuclear Information System (INIS)

    Williams, R.A.

    1975-01-01

    A dosimeter and method for use by personnel working in an area of mixed ionizing radiation fields for measuring and/or determining the effective energy of x- and gamma radiation; beta, x-, and gamma radiation dose equivalent to the surface of the body; beta, x-, and gamma radiation dose equivalent at a depth in the body; the presence of slow neutron, fast neutron dose equivalent; and orientation of the person wearing the dosimeter to the source of radiation is disclosed. Optionally integrated into this device and method are improved means for determining neutron energy spectrum and absorbed dose from fission gamma and neutron radiation resulting from accidental criticality

  14. The shelf life of dyed polymethylmethacrylate dosimeters

    International Nuclear Information System (INIS)

    Bett, R.; Watts, M.F.; Plested, M.E.

    2002-01-01

    The long-term stability of the radiation response of Harwell Red 4034 and Amber 3042 Perspex Dosimeters has been monitored for more than 15 years, and the resulting data used in the justification of their shelf-life specifications

  15. Indoor radon level measurements in Iran using AEOI passive dosimeters

    International Nuclear Information System (INIS)

    Sohrabi, M.; Solaymanian, A.R.

    1988-01-01

    A passive radon diffusion dosimeter was developed at the RPD of AEOI for nationwide indoor radon level measurements. Several parameters of the dosimeter were studied. Radon levels were determined in about 250 houses in Ramsar (a high natural radiation area), Tehran, Babolsar and Gonabad. In this paper, the results of some dosimeter parameters as well as radon levels in indoor air are reported

  16. Development of Thermoluminescence Dosimeter CaSO4:Dy as Personal and Environmental Dosimeters

    International Nuclear Information System (INIS)

    Hasnel Sofyan

    2009-01-01

    Development of personal and environmental dosimeters using material phosphors of CaSO 4 :Dy powder in form capillary glass and disc teflon thermoluminescence (TL) dosimeter have been done. TL dosimeter CaSO 4 :Dy powder used can record dose response less than 0.01 mGy. Fading of TL dosimeter capillary glass after 29 days is 25%. In 1 batch, making of CaSO 4 :Dy powder can obtain 2 groups of dosimeter capillaries with coefficient variance smaller than 10%. This discrepancy caused difference in powder making and reading of the TL dosimeter. TL dosimeter CaSO 4 :Dy teflon disc with dia. 5 mm and 0.8 mm thickness is homogeneous mixture between phosphor powder with dia. 80 to 150 mesh and teflon powder dia. 20 μm. The composition of CaSO 4 :Dy and teflon in TL dosimeter influence sensitivity of the dosimeter. It’ concluded that in order to obtain optimal sensitivity of TL dosimeter, the composition of CaSO 4 :Dy and teflon is 3 and 1 with pressured of disc in 700 MPa. (author).

  17. An improved dosimeter having constant flow pump

    International Nuclear Information System (INIS)

    Baker, W.B.

    1980-01-01

    A dosemeter designed for individual use which can be used to monitor toxic radon gas and toxic related products of radon gas in mines and which incorporates a constant air stream flowing through the dosimeter is described. (U.K.)

  18. SU-E-T-585: Optically-Stimulated Luminescent Dosimeters for Monitoring Pacemaker Dose in Radiation Therapy

    International Nuclear Information System (INIS)

    Apicello, L; Riegel, A; Jamshidi, A

    2015-01-01

    Purpose: A sufficient amount of ionizing radiation can cause failure to components of pacemakers. Studies have shown that permanent damage can occur after a dose of 10 Gy and minor damage to functionality occurs at doses as low as 2 Gy. Optically stimulated thermoluminescent dosimeters (OSLDs) can be used as in vivo dosimeters to predict dose to be deposited throughout the treatment. The purpose of this work is to determine the effectiveness of using OSLDs for in vivo dosimetry of pacemaker dose. Methods: As part of a clinical in vivo dosimetry experience, OSLDs were placed at the site of the pacemaker by the therapist for one fraction of the radiation treatment. OSLD measurements were extrapolated to the total dose to be received by the pacemaker during treatment. A total of 79 measurements were collected from November 2011 to December 2013 on six linacs. Sixty-six (66) patients treated in various anatomical sites had the dose of their pacemakers monitored. Results: Of the 79 measurements recorded, 76 measurements (96 %) were below 2 Gy. The mean and standard deviation were 50.12 ± 76.41 cGy. Of the 3 measurements that exceeded 2 Gy, 2 measurements matched the dose predicted in the treatment plan and 1 was repeated after an unexpectedly high Result. The repeated measurement yielded a total dose less than 2 Gy. Conclusion: This analysis suggests OSLDs may be used for in vivo monitoring of pacemaker dose. Further research should be performed to assess the effect of increased backscatter from the pacemaker device

  19. SU-E-T-585: Optically-Stimulated Luminescent Dosimeters for Monitoring Pacemaker Dose in Radiation Therapy

    Energy Technology Data Exchange (ETDEWEB)

    Apicello, L [Hofstra University, Hempstead, NY (United States); Riegel, A; Jamshidi, A [North Shore LIJ Health System, Lake Success, NY (United States)

    2015-06-15

    Purpose: A sufficient amount of ionizing radiation can cause failure to components of pacemakers. Studies have shown that permanent damage can occur after a dose of 10 Gy and minor damage to functionality occurs at doses as low as 2 Gy. Optically stimulated thermoluminescent dosimeters (OSLDs) can be used as in vivo dosimeters to predict dose to be deposited throughout the treatment. The purpose of this work is to determine the effectiveness of using OSLDs for in vivo dosimetry of pacemaker dose. Methods: As part of a clinical in vivo dosimetry experience, OSLDs were placed at the site of the pacemaker by the therapist for one fraction of the radiation treatment. OSLD measurements were extrapolated to the total dose to be received by the pacemaker during treatment. A total of 79 measurements were collected from November 2011 to December 2013 on six linacs. Sixty-six (66) patients treated in various anatomical sites had the dose of their pacemakers monitored. Results: Of the 79 measurements recorded, 76 measurements (96 %) were below 2 Gy. The mean and standard deviation were 50.12 ± 76.41 cGy. Of the 3 measurements that exceeded 2 Gy, 2 measurements matched the dose predicted in the treatment plan and 1 was repeated after an unexpectedly high Result. The repeated measurement yielded a total dose less than 2 Gy. Conclusion: This analysis suggests OSLDs may be used for in vivo monitoring of pacemaker dose. Further research should be performed to assess the effect of increased backscatter from the pacemaker device.

  20. Internal background build-up measurements in CaF2:Mn thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Balasybrahmanyam, V.; Measures, M.P.

    1977-01-01

    Some problems associated with the internal background build-up (IBB) of CaF 2 :Mn thermoluminescent dosimeters are reported. As a result of an investigation of batches of the EG and G model 15 dosimeter it is considered that measurements using this type of dosimeter are accurate and reproducible once the IBB has been determined. However, the use of the Manufacturer's claimed average of 0.064 mR/day can lead to erroneous results when determining environmental background dose rates. The authors therefore urge a rigid quality control program by the manufacturer and suggest that purchasers should be supplied with IBB information of each batch of dosimeters. Meanwhile each user should be aware of the IBB problem and be extremely cautious when using these dosimeters for environmental monitoring purposes. (U.K.)

  1. A quality control program for the thermoluminescence dosimeter (TLD) in personnel radiation monitoring

    International Nuclear Information System (INIS)

    Seo, Kyung Won; Kim, Jang Lyul; Lee, Sang Yoon; Lee, Hyung Sub

    1994-01-01

    High quality radiation dosimetry is essential for workers who rely upon personal dosemeters to record the amount of radiation to which they are exposed. The ministry of science and technology (MOST) issued a ministerial ordinance (No 199-15) about the technical criteria on personnel dosimeter processors on 1992. The purpose of this quality control program is to prescribe the procedures approved by the management of KAERI for implementing a quality badge service by means of TLD for personnel working in an area where they may be exposed to ionization radiation. (Author) 10 refs

  2. Evaluation of the implementation and use of active personal dosimeters for neutrons in Brazil

    International Nuclear Information System (INIS)

    Castro B, C. P.; Wagner P, W.; De Souza P, K. C.

    2014-08-01

    This work was conducted through of a field research based on a questionnaire sent to users of active personal dosimeters. A retrospective study of the last six years was also carried out of the services in the Neutron Metrology Laboratory (2008-2013) referent to the active personal dosimeters, taking into consideration the standards ISO-8529-3 and IEC-61526. The active personal dosimeters are defined as any instrument of individual monitoring with direct reading capacity, used by individuals exposed to ionizing radiation fields. Through research was verified that the active personal dosimeters work associated with other dosimeter types. Considering all dosimeters declared in the questionnaire, only two dosimeters (MGP brand Dmc 2000-GN model and the brand ATOMTEX model AT2503A) have conformity declaration with the international standard IEC-61526: 2005 reported by the manufacturers. (author)

  3. Reliability of an x-ray system for calibrating and testing personal radiation dosimeters

    Science.gov (United States)

    Guimarães, M. C.; Silva, C. R. E.; Rosado, P. H. G.; Cunha, P. G.; Da Silva, T. A.

    2018-03-01

    Metrology laboratories are expected to maintain standardized radiation beams and traceable standard dosimeters to provide reliable calibrations or testing of detectors. Results of the characterization of an x-ray system for performing calibration and testing of radiation dosimeters used for individual monitoring are shown in this work.

  4. Study on the angular dependence of personal exposure dosimeter - Focus on thermoluminescent dosimeter and photoluminescent dosimeter

    International Nuclear Information System (INIS)

    Dong, Kyung-Rae; Kweon, Dae Cheol; Chung, Woon-Kwan; Goo, Eun-Hoe; Dieter, Kevin; Choe, Chong-Hwan

    2011-01-01

    Radiation management departments place more emphasis on the accuracy of measurements than on the increase in the average dose and personal exposure dose from the use of radiation equipment and radioactive isotopes. Although current measurements are taken using devices, such as film badge dosimeters, pocket dosimeters and thermoluminescent dosimeters (TLDs), this study compared the angular dependence between the widely used TLDs and photoluminescent dosimeter (PLDs) in order to present primary data and evaluate the utility of PLD as a new dosimeter device. For X-ray fluoroscopy, a whole body phantom was placed on a table with a setting for the G-I technical factors fixed at a range of approximately 40 cm with a range of ±90 o at an interval scale of 15 o from the center location of an average radiological worker for PLDs (GD-450) and TLDs (Carot). This process was repeated 10 times, and at each time, the cumulative dosage was interpreted from 130 dosimeters using TLDs (UD-710R, Panasonic) and PLDs (FGD-650). The TLD and PLD showed a 52% and 23% decrease in the depth dosage from 0 o to -90 o , respectively. Therefore, PLDs have a lower angular dependence than TLDs.

  5. Procedure for the delivering of personal short-term visitor dosimeters

    CERN Document Server

    2016-01-01

    Update of the administrative procedure for delivering a personal short-term visitor dosimeter to associated members of CERN’s personnel.   Associated members of the CERN personnel may request a short term visitor dosimeter if working only in Supervised Radiation Areas and for a period of less than two months in a calendar year. Such a dosimeter is delivered without the need to provide the usual regular documents: radiation passport, certificate from the home institute or medical certificate. Periodic verification will ensure that holders of these personal dosimeters do not exceed the maximum allowed personal dose for this type of dosimeter, which is the same as the limit for members of the public at 1 mSv per year. From now on, the two-month period can be spread over a calendar year, offering greater flexibility to users coming to CERN for multiple short periods. Please return unused dosimeters Persons leaving CERN for a period of more than one month should return their dosimeter to the D...

  6. Offsite environmental monitoring report: Radiation monitoring around United States nuclear test areas, calendar year 1991

    Energy Technology Data Exchange (ETDEWEB)

    Chaloud, D.J.; Dicey, B.B.; Mullen, A.A.; Neale, A.C.; Sparks, A.R.; Fontana, C.A.; Carroll, L.D.; Phillips, W.G.; Smith, D.D.; Thome, D.J.

    1992-01-01

    This report describes the Offsite Radiation Safety Program conducted during 1991 by the Environmental Protection Agency`s (EPA`s) Environmental Monitoring Systems Laboratory-Las Vegas. This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ion chambers (PICs); and by biological monitoring of animals, food crops, and humans. Personnel with mobile monitoring equipment are placed in areas downwind from the test site prior to each nuclear weapons test to implement protective actions, provide immediate radiation monitoring, and obtain environmental samples rapidly after any occurrence of radioactivity release. Comparison of the measurements and sample analysis results with background levels and with appropriate standards and regulations indicated that there was no radioactivity detected offsite by the various EPA monitoring networks and no exposure above natural background to the population living in the vicinity of the NTS that could be attributed to current NTS activities. Annual and long-term trends were evaluated in the Noble Gas, Tritium, Milk Surveillance, Biomonitoring, TLD, PIC networks, and the Long-Term Hydrological Monitoring Program.

  7. Reliability of an x-ray system for calibrating and testing personal radiation dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Guimarães, M.C.; Silva, C.R.E.; Silva, T.A. da [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil); Rosado, P.H.G.; Cunha, P.G. [Instituto de Radioproteção e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil)

    2017-07-01

    Metrology laboratories are expected to maintain standardized radiation beams and traceable standard dosimeters to provide reliable calibrations or testing of detectors. Results of the characterization of an x-ray system for performing calibration and testing of radiation dosimeters used for individual monitoring are shown in this work. (author)

  8. Reliability of an x-ray system for calibrating and testing personal radiation dosimeters

    International Nuclear Information System (INIS)

    Guimarães, M.C.; Silva, C.R.E.; Silva, T.A. da; Rosado, P.H.G.; Cunha, P.G.

    2017-01-01

    Metrology laboratories are expected to maintain standardized radiation beams and traceable standard dosimeters to provide reliable calibrations or testing of detectors. Results of the characterization of an x-ray system for performing calibration and testing of radiation dosimeters used for individual monitoring are shown in this work. (author)

  9. Development of dark-striped field mice, Apodemus agrarius coreae, as a biological dosimeter in a radio-ecological monitoring system: 2. Survival rates and hematology

    International Nuclear Information System (INIS)

    Kim, Hee Sun; Kim, Chong Soon; Nishmura, Y.

    2005-01-01

    Regarding the management of nuclear power plants and the installation of facilities for radiation waste storage: social concerns over radiation safety are increasing. To understand how environmental radiation affects on human beings, the development of an reasonable monitoring system is required. The existing radio-environmental surveillance systems can be classified into physical and biological monitoring systems. The wild small animals and livestocks were reported to be effective biological indicators of environmental radiation This study investigated the possibility of using dark-striped field mice as a biological dosimetric model to assess the effect of radiation on the human environments. For this study, the criteria for the biological dosimeters of environmental radiation were established as the following: first, it should be an animal from a clear background of species; second, it should inhabit a broad range of areas and in considerable numbers; third, it should maintain identical ecological characteristics; fourth, it should be cohabitating with humans; fifth, it should have been consuming food found in their habitat; and finally, it should indicate a clear doseresponse relationship with high sensitivity. Based on such criteria, this study investigated the possibility of using dark-striped field mice as an effective biological dosimeter. Primarily, their species were classified based on their morphological external characteristics and isoenzymic patterns. The taxonomically classified darkstriped field mice, A. agrarius coreae, were then irradiated to investigate their radiation sensitivity based on the survival rate and hematology in this study

  10. Development of dark-striped field mice, Apodemus agrarius coreae, as a biological dosimeter in a radio-ecological monitoring system: 2. Survival rates and hematology

    Energy Technology Data Exchange (ETDEWEB)

    Kim, Hee Sun; Kim, Chong Soon [Korea Hydro and Nuclear Power Co., Ltd., Daejeon (Korea, Republic of); Nishmura, Y. [National Institute of Radiological Sciences, Chiba (Japan)

    2005-07-01

    Regarding the management of nuclear power plants and the installation of facilities for radiation waste storage: social concerns over radiation safety are increasing. To understand how environmental radiation affects on human beings, the development of an reasonable monitoring system is required. The existing radio-environmental surveillance systems can be classified into physical and biological monitoring systems. The wild small animals and livestocks were reported to be effective biological indicators of environmental radiation This study investigated the possibility of using dark-striped field mice as a biological dosimetric model to assess the effect of radiation on the human environments. For this study, the criteria for the biological dosimeters of environmental radiation were established as the following: first, it should be an animal from a clear background of species; second, it should inhabit a broad range of areas and in considerable numbers; third, it should maintain identical ecological characteristics; fourth, it should be cohabitating with humans; fifth, it should have been consuming food found in their habitat; and finally, it should indicate a clear doseresponse relationship with high sensitivity. Based on such criteria, this study investigated the possibility of using dark-striped field mice as an effective biological dosimeter. Primarily, their species were classified based on their morphological external characteristics and isoenzymic patterns. The taxonomically classified darkstriped field mice, A. agrarius coreae, were then irradiated to investigate their radiation sensitivity based on the survival rate and hematology in this study.

  11. Tissue-Equivalent Radiation Dosimeter-On-A-Chip, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — Available digital dosimeters are bulky and unable to provide real-time monitoring of dose for space radiation. The complexity of space-flight design requires...

  12. Automated Calibration of Dosimeters for Diagnostic Radiology

    International Nuclear Information System (INIS)

    Romero Acosta, A.; Gutierrez Lores, S.

    2015-01-01

    Calibration of dosimeters for diagnostic radiology includes current and charge measurements, which are often repetitive. However, these measurements are usually done using modern electrometers, which are equipped with an RS-232 interface that enables instrument control from a computer. This paper presents an automated system aimed to the measurements for the calibration of dosimeters used in diagnostic radiology. A software application was developed, in order to achieve the acquisition of the electric charge readings, measured values of the monitor chamber, calculation of the calibration coefficient and issue of a calibration certificate. A primary data record file is filled and stored in the computer hard disk. The calibration method used was calibration by substitution. With this system, a better control over the calibration process is achieved and the need for human intervention is reduced. the automated system will be used in the calibration of dosimeters for diagnostic radiology at the Cuban Secondary Standard Dosimetry Laboratory of the Center for Radiation Protection and Hygiene. (Author)

  13. Design, construction and characterization of a dosimeter for neutron radiation

    International Nuclear Information System (INIS)

    Souto, Eduardo de Brito

    2007-01-01

    An individual dosimeter for neutron-gamma mixed field dosimetry was design and developed aiming monitoring the increasing number of workers potentially exposed to neutrons. The proposed dosimeter was characterized to an Americium-Beryllium source spectrum and dose range of radiation protection interest (up to 20 mSv). Thermoluminescent albedo dosimetry and nuclear tracks dosimetry, traditional techniques found in the international literature, with materials of low cost and national production, were used. A commercial polycarbonate, named SS-1, was characterized for solid state tack detector application. The chemical etching parameters and the methodology of detectors evaluation were determined. The response of TLD-600, TLD-700 and SS-1 were studied and algorithms for dose calculation of neutron and gamma radiation of Americium- Beryllium sources were proposed. The ratio between thermal, albedo and fast neutrons responses, allows analyzing the spectrum to which the dosimeter was submitted and correcting the track detector response to variations in the radiation incidence angle. The new dosimeter is fully characterized, having sufficient performance to be applied as neutron dosimeter in Brazil. (author)

  14. Radiation dosimeter utilizing the thermoluminescence of lithium fluoride.

    Science.gov (United States)

    CAMERON, J R; DANIELS, F; JOHNSON, N; KENNEY, G

    1961-08-04

    A dosimeter, with little wavelength dependence and large useful energy range for electromagnetic radiation, which is simple to use and read, has been developed. It appears to have applications in personnel monitoring as well as radiation research.

  15. Personal noise dosimeters: Accuracy and reliability in varied settings

    Directory of Open Access Journals (Sweden)

    Sheri Lynn Cook-Cunningham

    2014-01-01

    during a day in the life of college music students. Three participants each wore two types of dosimeters for an 8-h period during a normal school day. Descriptive statistical analyzes including means, standard deviation and Pearson product-moment correlation. The primary finding is that the dosimeters in this study recorded results within ±2 dB of either a reference measurement or within dosimeters in all four conditions examined. All dosimeters studied measured steady noise source accurately and consistently, with strong positive correlations across all instruments. Measurements acquired during choral rehearsals indicated a maximum of 1.5 dB difference across dosimeters. The Etymotic research personal noise dosimeters (ER200D could provide individuals and schools of music with a relatively inexpensive tool to monitor sound doses. Findings from this study suggest that the three brands of dosimeters tested will provide reliable Leq levels and hearing dosages in both PN and natural settings.

  16. New junctionless RADFET dosimeter design for low-cost radiation monitoring applications

    International Nuclear Information System (INIS)

    Arar, Djemai; Djeffal, Faycal; Bentrcia, Toufik; Chahdi, Mohamed

    2014-01-01

    This paper is devoted to the presentation of a quantitative analysis of the Junctionless Gate All Around RADFET (JL GAA RADFET) dosimeter, where the numerical simulation has been carried out using the Atlas 3-D simulator. The impact of the total dose, alternative gate materials and the channel doping on the threshold voltage of the JL GAA RADFET is addressed. The obtained results have indicated a significant improvement in the subthreshold parameters when compared to the conventional GAA RADFET dosimeter. Therefore, the implementation of junctionless-based sensors in the near future can provide more accurate results with low costs, in addition to alleviating many difficulties in the measurement procedure. (copyright 2014 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  17. New junctionless RADFET dosimeter design for low-cost radiation monitoring applications

    Energy Technology Data Exchange (ETDEWEB)

    Arar, Djemai; Djeffal, Faycal [Department of Electronics, University of Batna, Batna 05000 (Algeria); Bentrcia, Toufik; Chahdi, Mohamed [Department of Physics, University of Batna, Batna 05000 (Algeria)

    2014-01-15

    This paper is devoted to the presentation of a quantitative analysis of the Junctionless Gate All Around RADFET (JL GAA RADFET) dosimeter, where the numerical simulation has been carried out using the Atlas 3-D simulator. The impact of the total dose, alternative gate materials and the channel doping on the threshold voltage of the JL GAA RADFET is addressed. The obtained results have indicated a significant improvement in the subthreshold parameters when compared to the conventional GAA RADFET dosimeter. Therefore, the implementation of junctionless-based sensors in the near future can provide more accurate results with low costs, in addition to alleviating many difficulties in the measurement procedure. (copyright 2014 WILEY-VCH Verlag GmbH and Co. KGaA, Weinheim) (orig.)

  18. Radiation monitoring around United States nuclear test areas, calendar year 1989

    International Nuclear Information System (INIS)

    1990-05-01

    This report describes the Offsite Radiation Safety Program conducted during 1989 by the Environmental Protection Agency's (EPA's) Environmental Monitoring Systems Laboratory-Las Vegas (EMSL-LV). This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels, and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether the testing is in compliance with existing radiation protection standards, and to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ion chambers (PICs); and by biological monitoring of both animals and humans. To implement protective actions, provide immediate radiation monitoring, and obtain environmental samples rapidly after any release of radioactivity, personnel with mobile monitoring equipment are placed in areas downwind from the test site prior to each test. Comparison of the measurements and sample analysis results with background levels and with appropriate standards and regulations indicated that there was no radioactivity detected offsite by the various EPA monitoring networks and no exposure above natural background to the population living in the vicinity of the NTS that could be attributed to NTS activities. Trends were evaluated in the Noble Gas and Tritium, Milk Surveillance, TLD, and PIC networks, and the Long-Term Hydrological Monitoring Program. 35 refs., 68 figs., 32 tabs

  19. Radiation monitoring around United States nuclear test areas, calendar year 1989

    Energy Technology Data Exchange (ETDEWEB)

    1990-05-01

    This report describes the Offsite Radiation Safety Program conducted during 1989 by the Environmental Protection Agency's (EPA's) Environmental Monitoring Systems Laboratory-Las Vegas (EMSL-LV). This laboratory operates an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels, and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether the testing is in compliance with existing radiation protection standards, and to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs) and using pressurized ion chambers (PICs); and by biological monitoring of both animals and humans. To implement protective actions, provide immediate radiation monitoring, and obtain environmental samples rapidly after any release of radioactivity, personnel with mobile monitoring equipment are placed in areas downwind from the test site prior to each test. Comparison of the measurements and sample analysis results with background levels and with appropriate standards and regulations indicated that there was no radioactivity detected offsite by the various EPA monitoring networks and no exposure above natural background to the population living in the vicinity of the NTS that could be attributed to NTS activities. Trends were evaluated in the Noble Gas and Tritium, Milk Surveillance, TLD, and PIC networks, and the Long-Term Hydrological Monitoring Program. 35 refs., 68 figs., 32 tabs.

  20. Individual dosimeter for radon and thoron daughters

    International Nuclear Information System (INIS)

    Chapuis, A.M.; Duport, P.; Zettwoog, P.

    1979-01-01

    The dosimeter is designed for the continuous measurement of the concentration of α emitters from the uranium 238 and thorium 232 series. It enables the measurement of, firstly the aerosol concentration of 218 Po (Radium A), 214 Po (Radium C') and 212 Po (Thorium C') and secondly the activity of long-lived α emitters in aerosols coming from ore dusts. One light weight version of this dosimeter is autonomous for 18 hours and is designed to measure individual doses, due to inhalation, for workers employed in uranium mines and ore processing plants. An other version using the same sampling head allows the monitoring of air concentrations in working environments. Living quarters, or free air

  1. Fundamentals of Polymer Gel Dosimeters

    Science.gov (United States)

    McAuley, Kim B.

    2006-12-01

    The recent literature on polymer gel dosimetry contains application papers and basic experimental studies involving polymethacrylic-acid-based and polyacrylamide-based gel dosimeters. The basic studies assess the relative merits of these two most commonly used dosimeters, and explore the effects of tetrakis hydroxymethyl phosphonium chloride (THPC) antioxidant on dosimeter performance. Polymer gel dosimeters that contain THPC or other oxygen scavengers are called normoxic dosimeters, because they can be prepared under normal atmospheric conditions, rather than in a glove box that excludes oxygen. In this review, an effort is made to explain some of the underlying chemical phenomena that affect dosimeter performance using THPC, and that lead to differences in behaviour between dosimeters made using the two types of monomer systems. Progress on the development of new more effective and less toxic dosimeters is also reported.

  2. Evaluation of a thermoluminescent dosimeter for personnel monitoring in the nuclear-radiation environment

    International Nuclear Information System (INIS)

    Johnson, F.A.; McGowan, S.; Gravelle, R.A.

    1983-09-01

    Efforts to improve the responses of phosphor/polyethylene TLDs to neutrons in the 1-MeV range have been only partially successful. Some increases in response have resulted from reduction of phosphor grain size and substitution of CaSO 4 :Tm with CaF 2 :Mn or CaF 2 :Dy. Reasons for the continuing low response are not fully understood. Calculated responses of the TLDs, when located on the body, show a closer correlatin with the dose received by the bone marrow as a result of exposure to a weapon spectrum than is evident for the case of flat-response dosimeters. However the TLDs exhibit a greater difference in their response to a weapon spectrum as compared with that to a pure gamma spectrum than is the case for a flat-response dosimeter. The disparity becomes even larger when the dose measured by the dosimeters is expressed in terms of the biologically effective dose absorbed in the bone marrow. Investigation of the ability of a dosimeter, which is sensitive only to γ-rays, to provide a measure of the total bone marrow dose produced by a combined flux of γ-rays and neutrons, indicates that there is an unacceptable difference between the response to a pure gamma spectrum and that to a mixed spectrum from a weapon

  3. Real-time personal dose monitoring and management system

    International Nuclear Information System (INIS)

    Zhang Zhiyong; Cheng Chang; Yang Huating; Liu Zhengshan; Deng Changming; Li Mei

    2000-01-01

    This paper mainly describes a real-time personal dose monitoring and management system. The system is composed of three parts that include SDM-98 semiconductor detector personal dosimeters, Data Readers and a Management System Software. It can be used for personal dose monitoring and management and other controlling actions in a radioactive controlled area. Adopting semiconductor detector and microcontroller, SDM-98 Personal Dosimeter is used to measure personal accumulated dose equivalent and dose rate caused by X-ray and Gamma ray. The results can be read directly on LCD. All the data stored in dosimeter can be transmitted into a data reader by infrared optical link. The alarm threshold can be adjusted successively in whole range of dose or dose rate. The Data Reader is an intelligent interface between the dosimeter and master computer. The data received from dosimeter will be sent to a master computer through RS-232 serial interface. According to the master computer's order, the Data Reader can turn on the dosimeter's power at entrance and shutdown it at exit. The Management System Software which written by Visual BASIC 5.0 runs on MS Win95. All the measuring data from dosimeters can be analyzed and treated according to requirements and stored in database. Therefore, some figures and tables relative to dose or rate can be shown on screen or printed out. (author)

  4. Thermoluminescence and phosphate glass dosimeter systems in the low dose range

    International Nuclear Information System (INIS)

    Piesch, E.; Burgkhardt, B.

    1978-06-01

    This report describes a standard test program for TLD and RPL systems worked out by the Working Party on 'Dose Measurement of External Radiation' by the Fachverband fuer Strahlenschutz e.V. to demonstrate the performance of dosimeter systems to be employed in environmental monitoring and in personnel dosimetry. The results of an intercomparison study are outlined in which 17 laboratories from the German speaking countries participated with 43 dosimeter systems. (orig.) [de

  5. Dosimeter design specifications

    International Nuclear Information System (INIS)

    Anon.

    1981-01-01

    The combination dosimeter and security credential holder was developed as part of the effort involved to provide an automated readout and thermoluminescent dosimetry capability at Hanford. The holder is designed to accomodate the thermoluminescent dosimeter card, appropriate filters, the security credential and a snap type clip. The body of the holder is ABS plastic (acrylontrile-butadiene-styrene). The dosimeter holder and card is mold casted providing uniformity of construction

  6. Estimation of personal dose based on the dependent calibration of personal dosimeters in interventional radiology

    International Nuclear Information System (INIS)

    Mori, Hiroshige; Koshida, Kichiro; Ichikawa, Katsuhiro

    2007-01-01

    The purpose of present study is, in interventional radiology (IVR), to elucidate the differences between each personal dosimeter, and the dependences and calibrations of area or personal dose by measurement with electronic dosimeters in particular. We compare space dose rate distributions measured by an ionization survey meter with the value measured by personal dosimeter: an optically stimulated luminescence, two fluoroglass, and two electronic dosimeters. Furthermore, with electronic dosimeters, we first measured dose rate, energy, and directional dependences. Secondly, we calibrated the dose rate measured by electronic dosimeters with the results, and estimated these methods with coefficient of determination and Akaike's Information Criterion (AIC). The results, especially in electronic dosimeters, revealed that the dose rate measured fell by energy and directional dependences. In terms of methods of calibration, the method is sufficient for energy dependence, but not for directional dependence, because of the lack of stable calibration. This improvement poses a question for the future. The study suggested that these dependences of the personal dosimeter must be considered when area or personal dose is estimated in IVR. (author)

  7. X-rays individual dose assessment using TLD dosimeters

    International Nuclear Information System (INIS)

    Salas, Carlos

    2008-01-01

    This paper describes the methodology used in Embalse NPP for measuring individual X-ray dose in dentists and radiologists, who work in areas near the plant. Personnel is provided with TLD personal dosimeters for thoracic use, as well as TLD ring dosimeters. This individual X-ray dosimetry is fundamental in order to know the effective energy coming from the radiation field, since the dosimetry factors depend on it. On the other hand, the response of the TLD crystals also depends of the effective energy; this accentuates the problem when assessing the individual dose. The X-ray dosimeter must simultaneously determine the value of the effective energy and the corresponding dose value. The basic principle for determining effective energy is by using at least two different TLD materials covered by filters of different thickness. The TLD materials used have totally energy responses. Therefore, different readouts from each of the crystals are obtained. The ratio between both readouts provides a factor that depends of the effective energy but that is 'independent' from the exposure values irradiated to the dosimeter. The Personal TLD dosimeter currently in use is Bicron-Harshaw. It comprises a carrier model 8807. This carrier contains a card model 2211 which groups two TLD 200 crystals and two TLD 100 crystals. It has internal filters at each side of the TLD 200 crystals. The periodical calibration of these dosimeters consists in the irradiation of some dosimeters with different X-ray energy beams in the National Atomic Energy Commission (CNEA). This dosimeter was used, by the National Regulatory Authority (ARN) in several comparisons, always getting satisfactory results. (author)

  8. The radiation dosimeter on-board the FY-4 Satellite

    Science.gov (United States)

    Zhang, B.; Sun, Y.; Zhang, S.; Zhang, X.; Sun, Y.; Jing, T.

    2017-12-01

    The total radiation dose effect can lead to a decrease in the performance of satellite devices or materials. Accurately obtaining the total radiation dose during satellite operation could help to analyze the abnormality of payloads in orbit and optimize the design of radiation shielding. The radiation dosimeter is one of the space environmental monitoring devices on the "FY-4" satellite, which is a new generation of geostationary meteorological satellite. The dosimeter consists of 8 detectors, which are installed in different locations of the satellite, to obtain the total radiation dose with different shielding thickness and different orientations. To measure a total radiation dose up to 2000krad(Si), 100nm ion implantation RADFET was used. To improve the sensitivity of the dosimeter, the bias voltage of RADFET is set to 15V, and a 10V, 15-bit A/D is adopted to digitalize the RADFET's threshold voltage, which is increased as the total radiation dose grows. In addition, the temperature effect of RADFET is corrected from the measured temperature on orbit. The preliminary monitoring results show that the radiation dose is less than 35rad (Si) per day at 0.87 mm shielding thickness of equivalent aluminum in the geostationary orbit, and the dose in Y direction of the satellite is less than those in the X and Z directions. The radiation dose at the thickness of 3.87 mm equivalent aluminum is less than 1rad(Si)/day. It is found that the daily total dose measured by the dosimeter has a strong correlation with the flux of high energy electrons.

  9. Offsite environmental monitoring report; radiation monitoring around United States nuclear test areas, Calendar Year 1996

    International Nuclear Information System (INIS)

    Davis, M.G.; Flotard, R.D.; Fontana, C.A.; Huff, P.A.; Maunu, H.K.; Mouck, T.L.; Mullen, A.A.; Sells, M.D.

    1997-08-01

    This report describes the Offsite Radiation Safety Program. This laboratory operated an environmental radiation monitoring program in the region surrounding the Nevada Test Site (NTS) and at former test sites in Alaska, Colorado, Mississippi, Nevada, and New Mexico. The surveillance program is designed to measure levels and trends of radioactivity, if present, in the environment surrounding testing areas to ascertain whether current radiation levels and associated doses to the general public are in compliance with existing radiation protection standards. The surveillance program additionally has the responsibility to take action to protect the health and well being of the public in the event of any accidental release of radioactive contaminants. Offsite levels of radiation and radioactivity are assessed by sampling milk, water, and air; by deploying thermoluminescent dosimeters (TLDs); and using pressurized ionization chambers (PICs). No nuclear weapons testing was conducted in 1996 due to the continuing nuclear test moratorium. During this period, R and IE personnel maintained readiness capability to provide direct monitoring support if testing were to be resumed and ascertained compliance with applicable EPA, DOE, state, and federal regulations and guidelines. Comparison of the measurements and sample analysis results with background levels and with appropriate standards and regulations indicated that there was no airborne radioactivity from diffusion or resuspension detected by the various EPA monitoring networks surrounding the NTS. There was no indication of potential migration of radioactivity to the offsite area through groundwater and no radiation exposure above natural background was received by the offsite population. All evaluated data were consistent with previous data history

  10. Assessing Doses to Interventional Radiologists Using a Personal Dosimeter Worn Over a Protective Apron

    Energy Technology Data Exchange (ETDEWEB)

    Stranden, E.; Widmark, A.; Sekse, T. (Buskerud Univ. College, Drammen (Norway))

    2008-05-15

    Background: Interventional radiologists receive significant radiation doses, and it is important to have simple methods for routine monitoring of their exposure. Purpose: To evaluate the usefulness of a dosimeter worn outside the protective apron for assessments of dose to interventional radiologists. Material and Methods: Assessments of effective dose versus dose to dosimeters worn outside the protective apron were achieved by phantom measurements. Doses outside and under the apron were assessed by phantom measurements and measurements on eight radiologists wearing two routine dosimeters for a 2-month period during ordinary working conditions. Finger doses for the same radiologists were recorded using thermoluminescent dosimeters (TLD; DXT-RAD Extremity dosimeters). Results: Typical values for the ratio between effective dose and dosimeter dose were found to be about 0.02 when the radiologist used a thyroid shield and about 0.03 without. The ratio between the dose to the dosimeter under and outside a protective apron was found to be less than 0.04. There was very good correlation between finger dose and dosimeter dose. Conclusion: A personal dosimeter worn outside a protective apron is a good screening device for dose to the eyes and fingers as well as for effective dose, even though the effective dose is grossly overestimated. Relatively high dose to the fingers and eyes remains undetected by a dosimeter worn under the apron

  11. Assessing Doses to Interventional Radiologists Using a Personal Dosimeter Worn Over a Protective Apron

    International Nuclear Information System (INIS)

    Stranden, E.; Widmark, A.; Sekse, T.

    2008-01-01

    Background: Interventional radiologists receive significant radiation doses, and it is important to have simple methods for routine monitoring of their exposure. Purpose: To evaluate the usefulness of a dosimeter worn outside the protective apron for assessments of dose to interventional radiologists. Material and Methods: Assessments of effective dose versus dose to dosimeters worn outside the protective apron were achieved by phantom measurements. Doses outside and under the apron were assessed by phantom measurements and measurements on eight radiologists wearing two routine dosimeters for a 2-month period during ordinary working conditions. Finger doses for the same radiologists were recorded using thermoluminescent dosimeters (TLD; DXT-RAD Extremity dosimeters). Results: Typical values for the ratio between effective dose and dosimeter dose were found to be about 0.02 when the radiologist used a thyroid shield and about 0.03 without. The ratio between the dose to the dosimeter under and outside a protective apron was found to be less than 0.04. There was very good correlation between finger dose and dosimeter dose. Conclusion: A personal dosimeter worn outside a protective apron is a good screening device for dose to the eyes and fingers as well as for effective dose, even though the effective dose is grossly overestimated. Relatively high dose to the fingers and eyes remains undetected by a dosimeter worn under the apron

  12. Characterization of the personal dosimeter Rn-disk for monitoring radon exposure

    International Nuclear Information System (INIS)

    Orlando, P.; Arcovito, G.; Amici, M.; Orlando, C.; Cardellini, F.; Fiorentino, A.; Trevisi, R.

    2009-01-01

    Rn-disk is a new passive device for measuring occupational exposure to radon 222, are presented the results of tests for the characterization of the dosimeter as a tool for estimating the individual dose for workers. [it

  13. Development of colored alumilite dosimeter

    CERN Document Server

    Obara, K; Yagi, T; Yokoo, N

    2003-01-01

    In the ITER (International Thermonuclear Experimental Reactor), in-vessel components such as blanket and divertor, which are installed in the vacuum vessel of the ITER, are maintained by remote handling equipment (RH equipment). The RH equipment for maintenance is operated under sever environmental conditions, such as high temperature (50 approx 100 degC), high gamma-ray radiation (approx 1 kGy/h) in an atmosphere of inert gas or vacuum; therefore many components of the RH equipment must have a suitable radiation resistance efficiency for long time operation (10 approx 100 MGy). Typical components of the RH equipment have been extensively tested under an intensive gamma-ray radiation. Monitoring of the radiation dose of the components of the RH equipment is essential to control the operation period of the RH equipment considering radiation resistance. However, the maximum measurable radiation dose of the conventional dosimeters, such as ionization chamber, liquid, glass and plastic dosimeters are limited to b...

  14. Passive radon daughter dosimeters

    International Nuclear Information System (INIS)

    McElroy, R.G.C.; Johnson, J.R.

    1986-03-01

    On the basis of an extensive review of the recent literature concerning passive radon daughter dosimeters, we have reached the following conclusions: 1) Passive dosimeters for measuring radon are available and reliable. 2) There does not presently exist an acceptable passive dosimeter for radon daughters. There is little if any hope for the development of such a device in the foreseeable future. 3) We are pessimistic about the potential of 'semi-passive dosimeters' but are less firm about stating categorically that these devices cannot be developed into a useful radon daughter dosimeter. This report documents and justifies these conclusions. It does not address the question of the worker's acceptance of these devices because at the present time, no device is sufficiently advanced for this question to be meaningful. 118 refs

  15. The Small Mixed Field Autonomous Radiation Tracker (SMART) Dosimeter, Phase I

    Data.gov (United States)

    National Aeronautics and Space Administration — Active dosimeters for astronauts and space weather monitors are critical tools for mitigating radiation induced health issues or system failure on capital equipment....

  16. Dosimeter charging and/or reading apparatus

    International Nuclear Information System (INIS)

    Fine, L.T.; Jackson, T.P.

    1980-01-01

    A device is disclosed for charging and/or reading a capacitor associated with an electrometer incorporated in a radiation dosimeter for the purpose of initializing or ''zeroing'', the dosimeter at the commencement of a radiation measurement cycle or reading it at any time thereafter. The dosimeter electrometer has a movable electrode the position of which is indicative of the charge remaining on the dosimeter capacitor and in turn the amount of radiation incident on the dosimeter since it was zeroed. The charging device also includes means for discharging, immediately upon conclusion of the dosimeter capacitor charging operation, stray capacitance inherent in the dosimeter by reason of its mechanical construction. The charge on the stray capacitance, if not discharged at the conclusion of the dosimeter capacitor charging operation, leaks off during the measurement cycle, introducing measurement errors. A light source and suitable switch means are provided for automatically illuminating the movable electrode of the dosimeter electrometer as an incident to charging the dosimeter capacitor to facilitate reading the initial, or ''zero'', position of the movable electrometer electrode after the dosimeter capacitor has been charged and the stray capacitance discharged. Also included is a manually actuatable switch means, which is operable independently of the aforementioned automatic switch means, to energize the lamp and facilitate reading of the dosimeter without charging

  17. Comparison between dosimeter films and electronic dosimeters results obtained in F-18 production practices at IEN/CNEN

    International Nuclear Information System (INIS)

    Paulo, Osvaldir; Carlos, Luiz R.J.; Mendes, Milton

    2008-01-01

    Full text: The aim of the present work is the comparison of the dose rates values obtained from dosimeter films with the values obtained from electronic dosimeters used by radiological protection technician involved in practices with fluorine-18 production in CV-28 cyclotron at IEN/CNEN. The motivation for this work was the increase on the demand of this radiopharmaceutical. Nowadays there is an increase demand on the number of clinics and hospitals which had started to use the technique of radiodiagnostic using positron-emission tomography (PET). Therefore, the produced radionuclide activities (physics quantity) and also the complexity of the practices of radiological protection in the areas of that facility had increased a lot. The conventional statistical methods will be used to evaluate the results obtained in the analysis of the data obtained in the reading of the dosimeters. This comparative method is however a first step to taking decisions regarding radiological protection service. Therefore that the radiometric surveys of routine are taken in consideration, as well as the information from the technician of radiological protection that is working in those areas. All this data will give support to improve and implement methods and practices on the accelerators' facilities. In addition, the use of these electronic dosimeters will makes possible the technician read these dose rates received immediately during the practices and use these information you make decisions. In case of dose rates values to present above the expected the practice will be ploughed. Another point to be considered is that this work will propitiate the future updates of the procedures designated to those practices and to conduct at the radiological protection optimization. (author)

  18. Light scattering in optical CT scanning of Presage dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Y; Adamovics, J; Cheeseborough, J C; Chao, K S; Wuu, C S, E-mail: yx2010@columbia.ed

    2010-11-01

    The intensity of the scattered light from the Presage dosimeters was measured using a Thorlabs PM100D optical power meter (Thorlabs Inc, Newton, NJ) with an optical sensor of 1 mm diameter sensitive area. Five Presage dosimeters were made as cylinders of 15.2 cm, 10 cm, 4 cm diameters and irradiated with 6 MV photons using a Varian Clinac 2100EX. Each dosimeter was put into the scanning tank of an OCTOPUS' optical CT scanner (MGS Research Inc, Madison, CT) filled with a refractive index matching liquid. A laser diode was positioned at one side of the water tank to generate a stationary laser beam of 0.8 mm width. On the other side of the tank, an in-house manufactured positioning system was used to move the optical sensor in the direction perpendicular to the outgoing laser beam from the dosimeters at an increment of 1 mm. The amount of scattered photons was found to be more than 1% of the primary light signal within 2 mm from the laser beam but decreases sharply with increasing off-axis distance. The intensity of the scattered light increases with increasing light attenuations and/or absorptions in the dosimeters. The scattered light at the same off-axis distance was weaker for dosimeters of larger diameters and for larger detector-to-dosimeter distances. Methods for minimizing the effect of the light scattering in different types of optical CT scanners are discussed.

  19. Electronic dosimeter characteristics and new developments

    International Nuclear Information System (INIS)

    Thompson, I.M.G.

    1999-01-01

    Electronic dosimeters are very much more versatile than existing passive dosimeters such as TLDs and film badges which have previously been the only type of dosimeters approved by national authorities for the legal measurement of doses to occupationally exposed workers. Requirements for the specifications and testing of electronic dosimeters are given in the standards produced by the International Electrotechnical Commission Working Group IEC SC45B/B8. A description is given of these standards and the use of electronic dosimeters as legal dosimeters is discussed. (author)

  20. Comparison of alanine dosimeters using silicone as their binder to a commercial, polystyrene-bound, alanine dosimeter

    International Nuclear Information System (INIS)

    Galindo, S.; Urena-Nunez, F.

    1997-01-01

    The feasibility of practical boron-containing alanine ESR dosimeters for gamma-neutron mixed field irradiation dosimeters depends in part on whether the γ response characteristics of these silicone-bound dosimeters are comparable to those of a commercially available dosimeter that has been used by the International Atomic Energy Agency (International Dose Assurance Service) as a transfer reference dosimeter. This work presents the results of the comparison of 3 batches of silicone-bound alanine dosimeters. The first batch consists of a mixture of alanine and boric acid; the second, alanine and borax; and the last contains only alanine. Results indicate that γ response characteristics of the silicone-bound samples are comparable to those of the commercial, polystyrene-bound, alanine dosimeter and that silicone has a strong potential as a binding substance for alanine ESR dosimetry. (Author)

  1. Development of a TL personal dosimeter identifiable PA exposure, and comparison with commercial TL dosimeters

    International Nuclear Information System (INIS)

    Kwon, J.W.; Kim, H.K.; Lee, J.K.; Kim, J.L.

    2004-01-01

    A single-dosimeter worn on the anterior surface of the body of a worker was found to significantly underestimate the effective dose to the worker when the radiation comes from the back. Several researchers suggested that this sort of underestimation can be corrected to a certain extent by using an extra dosimeter on the back. However, use of multiple dosimeters also has disadvantages such as complication in control or incurrence of extra cost. Instead of the common multi-dosimeter approach, in this study, a single dosimeter introducing asymmetric filters which enabled to identify PA exposure was designed, and its dose evaluation algorithm for AP-PA mixed radiation fields was established. A prototype TL personal dosimeter was designed and constructed. The Monte Carlo simulations were utilized in the design process and verified by experiments. The dosimeter and algorithm were applicable to photon radiation having an effective energy beyond 100 keV in AP-PA mixed radiation fields. A simplified performance test based on ANSI N13.11 showed satisfactory results. Considering that the requirements of the International Electrotechnical Commission (IEC) and the American National Standards Institute (ANSI) with regard to the dosimeter on angular dependency is reinforced, the dosimeter and the dose evaluation algorithm developed in this study provides a useful approach in practical personal dosimetry against inhomogeneous high energy radiation fields. (author)

  2. A method for reducing energy dependence of thermoluminescence dosimeter response by means of filters

    International Nuclear Information System (INIS)

    Bapat, V.N.

    1980-01-01

    This work describes the application of the method of partial surface shielding for reducing the energy dependence of the X-ray and γ-ray response of a dosimeter containing a CaSO 4 :Dy thermoluminescent phosphor mixed with KCl. in pellet form. Results are given of approximate computation of filter combinations that accomplish this aim, and of experimental verifications. Incorporation of the described filter combination makes it possible to use this relatively sensitive dosimeter for environmental radiation monitoring. A similar approach could be applied to any type of dosimeter in the form of a thin pellet or wafer. (author)

  3. Systematic approach to personnel neutron monitoring

    International Nuclear Information System (INIS)

    Griffith, R.V.; Hankins, D.E.

    1980-01-01

    NTA film and albedo detectors represent the major portion of personnel dosimeters now used for occupational neutron monitoring. However, recent attention to the spectral response of these systems has demonstrated the need for detectors that have a better match to the fields being monitored. Recent developments in direct recoil track etch dosimeters present some intriguing alternatives, and careful use of 237 Np fission fragment detectors offers the advantage of a good dose equivalent spectral match. Work continues on a number of other new detector mechanisms, but problems with sensitivity, energy response, gamma interference, etc., continue to prevent development of most mechanisms into viable personnel dosimeters. Current dosimeter limitations make a systematic approach to personnel neutron monitoring particularly important. Techniques have been developed and tested, using available portable survey instruments, that significantly improve the quality of dosimeter interpretation. Even simple spectrometry can be done with modest effort, significantly improving the health physicists ability to provide accurate neutron monitoring

  4. Locations of criticality alarms and nuclear accident dosimeters at Hanford

    International Nuclear Information System (INIS)

    1992-08-01

    Hanford facilities that contain fissionable materials capable of achieving critical mass are monitored with nuclear accident dosimeters (NADS) in compliance with the requirements of DOE Order 5480.11, Chapter XI, Section 4.c. (DOE 1988). The US Department of Energy (DOE) Richland Field Office (RL) has assigned the responsibility for maintaining and evaluating the Hanford NAD system to the Instrumentation and External Dosimetry (I ampersand ED) Section of Pacific Northwest Laboratory's (PNL's) Health Physics Department. This manual provides a description of the Hanford NAD, criteria and instructions for proper NAD placement, and the locations of these dosimeters onsite

  5. Selfcalibrated alanine/EPR dosimeters. A new generation of solid state/EPR dosimeters

    International Nuclear Information System (INIS)

    Yordanov, N.D.; Gancheva, V.

    1999-01-01

    Alanine/EPR dosimeters are well established as secondary, reference dosimeters for high-energy radiation. However, there are various sources of uncertainty in the evaluation of absorbed dose. This arises primarily from the necessity to calibrate each EPR spectrometer and each batch of dosimeters before their use. In order to overcome this disadvantage, a new generation alanine/EPR dosimeter has been developed, and its possibilities as a radiation detector are reported. Principally, it is a mixture of alanine, some quantity of EPR active substance, and a binding material. The EPR active substance, acting as an internal EPR standard, is chosen to have EPR parameters which are independent of the irradiation dose. The simultaneous recording of the spectra of both the sample and the standard under the same experimental conditions and the estimation of the ratio I alanine /I Mn as a function of the absorbed dose strongly reduces the uncertainties. The response of these dosimeters for 60 Co γ-radiation exhibits excellent linearity and reproducibility in the range of absorbed dose, 10 2 - 5 x 10 4 Gy. (author)

  6. Heater design for reading radiation dosimeters

    International Nuclear Information System (INIS)

    Seidel, J.G.; Felice, P.E.

    1982-01-01

    The nichrome heating element of a conventional dosimeter reading apparatus has been redesigned to include a flat-bottomed depression big enough to hold a thermoluminescent dosimeter. A thin glass plate is positioned in the recess on top of the dosimeter to retain it in the recess during the heating and reading process. This technique of securing the dosimeter in contact with the heating element avoids physical scratching or damage to the dosimeter

  7. A personal radio-frequency dosimeter with cumulative-dose recording capabilities

    International Nuclear Information System (INIS)

    Rochelle, R.W.; Moore, M.R.; Thomas, R.S.; Ewing, P.D.; Hess, R.A.; Hoffheins, B.S.

    1990-01-01

    The radio-frequency (rf) dosimeter developed by the Oak Ridge National Laboratory is a portable, pocket-sized cumulative-dose recording device designed to detect and record the strengths and durations of electric fields present in the work areas of naval vessels. The device measures an integrated dose and records the electric fields that exceed the permissible levels set by the American National Standards Institute. Features of the rf dosimeter include a frequency range of 30 MHz to 10 GHz and a three-dimensional sensor. Data obtained with the rf dosimeter will be used to determine the ambient field-strength profile for shipboard personnel over an extended time. Readings are acquired and averaged over a 6-min period corresponding to the rise time of the core body temperature. These values are stored for up to 6 months, after which the data are transferred to a computer via the dosimeter's serial port. The rf dosimeter should increase knowledge of the levels of electric fields to which individuals are exposed. 13 refs., 16 figs., 2 tabs

  8. Photon energy response of an aluminum oxide TLD environmental dosimeter

    International Nuclear Information System (INIS)

    Olsher, R.H.

    1992-01-01

    Because of aluminum oxide's significant advantage in sensitivity (about a factor of 30) over LiF, minimal fading characteristics and ease of processing, aluminum oxide thermoluminescent dosimeters (TLDS) are being phased in at Los alamos for environmental monitoring of photon radiation. The new environmental dosimeter design consists of a polyethylene holder, about 0. 5 cm thick, loaded with a stack of four aluminum oxide TLD chips, each 1 mm thick and 5 mm in diameter. As part of the initial evaluation of the new design, the photon energy response of the dosimeter was calculated over the range from 10 keV to 1 MeV. Specific goals of the analysis included the determination of individual chip response in the stack, assessment of the response variation due to TLD material (i.e., LiF versus A1 2 O 3 ), and the effect of copper filtration in flattening the response

  9. Personnel radiation dosimetry laboratory accreditation programme for thermoluminescent dosimeters : a proposal

    International Nuclear Information System (INIS)

    Bhatt, B.C.; Srivastava, J.K.; Iyer, P.S.; Venkatraman, G.

    1993-01-01

    Accreditation for thermoluminescent dosimeters is the process of evaluating a programme intending to use TL personnel dosimeters to measure, report and record dose equivalents received by radiation workers. In order to test the technical competence for conducting personnel dosimetry service as well as to decentralize personnel monitoring service, it has been proposed by Radiological Physics Division (RPhD) to accredit some of the laboratories, in the country. The objectives of this accreditation programme are: (i) to give recognition to competent dosimetry processors, and (ii) to provide periodic evaluation of dosimetry processors, including review of internal quality assurance programme to improve the quality of personnel dosimetry processing. The scientific support for the accreditation programme will be provided by the scientific staff from Radiological Physics Division (RPhD) and Radiation Protection Services Division (RPSD). This paper describes operational and technical requirements for the Personnel Radiation Dosimetry Laboratory Accreditation Programme for Thermoluminescent Dosimeters for Personnel Dosimetry Processors. Besides, many technical documents dealing with the TL Personnel Dosimeter System have been prepared. (author). 5 refs., 2 figs

  10. Hanford personnel dosimeter supporting studies FY-1981

    International Nuclear Information System (INIS)

    1982-08-01

    This report examined specific functional components of the routine external personnel dosimeter program at Hanford. Components studied included: dosimeter readout; dosimeter calibration; dosimeter field response; dose calibration algorithm; dosimeter design; and TLD chip acceptance procedures. Additional information is also presented regarding the dosimeter response to light- and medium-filtered x-rays, high energy photons and neutrons. This study was conducted to clarify certain data obtained during the FY-1980 studies

  11. Radiation dosimeters for medical use

    International Nuclear Information System (INIS)

    Risticj, S. Goran

    2013-01-01

    The several personal radiation dosimeter types for medical use, which look like promising for this kind of application, as pMOS (RADFET) dosimeter, direct ion storage (DIS) dosimeters, thermoluminescent (TL) and optically stimulated luminescent (OSL) dosimeters, are described, and their advantages and disadvantages are analyzed. The p-channel metal-oxide-semiconductor (pMOS) dosimetric transistors allow dose measurements in vivo in real time, and they are especially important for radiotherapy. Direct ion storage (DIS) dosimeters are a hybrid of ion chamber and floating gate MOSFETs (FGMOSFETs), show very high sensitivity. Radiative processes that happen during the exposure of crystal to radiation are classified as prompt luminescence or radioluminescence (RL). In the case of an emission during stimulation, this phenomenon is referred to thermoluminescence or optically stimulated luminescence depending on whether the stimulation source is heat or light. TL and OSL dosimeters are natural or synthetic materials, which the intensity of emitted light is proportional to the irradiation dose. (Author)

  12. Wireless remote radiation monitoring system (WRRMS). Innovative technology summary report

    International Nuclear Information System (INIS)

    1998-12-01

    The Science Application International Corporation (SAIC) RadStar trademark wireless remote radiation monitoring system (WRRMS) is designed to provide real-time monitoring of the radiation dose to workers as they perform work in radiologically contaminated areas. WRRMS can also monitor dose rates in a room or area. The system uses radio-frequency communications to transmit dose readings from the wireless dosimeters worn by workers to a remote monitoring station that can be located out of the contaminated area. Each base station can monitor up to 16 workers simultaneously. The WRRMS can be preset to trigger both audible and visual alarms at certain dose rates. The alarms are provided to the worker as well as the base station operator. This system is particularly useful when workers are wearing personal protective clothing or respirators that make visual observation of their self-reading dosimeters (SRDs), which are typically used to monitor workers, more difficult. The base station is an IBM-compatible personal computer that updates and records information on individual workers every ten seconds. Although the equipment costs for this improved technology are higher than the SRDs (amortized at $2.54/hr versus $1.02/hr), total operational costs are actually less ($639/day versus $851/day). This is because the WRRMS requires fewer workers to be in the contaminated zone than the traditional (baseline) technology. There are also intangible benefits associated with improved worker safety and as low as reasonably achievable (ALARA) principles, making the WRRMS an attractive alternative to the baseline technology. The baseline technology measures only integrated dose and requires workers to check their own dosimeters manually during the task

  13. Wireless remote radiation monitoring system (WRRMS). Innovative technology summary report

    Energy Technology Data Exchange (ETDEWEB)

    1998-12-01

    The Science Application International Corporation (SAIC) RadStar{trademark} wireless remote radiation monitoring system (WRRMS) is designed to provide real-time monitoring of the radiation dose to workers as they perform work in radiologically contaminated areas. WRRMS can also monitor dose rates in a room or area. The system uses radio-frequency communications to transmit dose readings from the wireless dosimeters worn by workers to a remote monitoring station that can be located out of the contaminated area. Each base station can monitor up to 16 workers simultaneously. The WRRMS can be preset to trigger both audible and visual alarms at certain dose rates. The alarms are provided to the worker as well as the base station operator. This system is particularly useful when workers are wearing personal protective clothing or respirators that make visual observation of their self-reading dosimeters (SRDs), which are typically used to monitor workers, more difficult. The base station is an IBM-compatible personal computer that updates and records information on individual workers every ten seconds. Although the equipment costs for this improved technology are higher than the SRDs (amortized at $2.54/hr versus $1.02/hr), total operational costs are actually less ($639/day versus $851/day). This is because the WRRMS requires fewer workers to be in the contaminated zone than the traditional (baseline) technology. There are also intangible benefits associated with improved worker safety and as low as reasonably achievable (ALARA) principles, making the WRRMS an attractive alternative to the baseline technology. The baseline technology measures only integrated dose and requires workers to check their own dosimeters manually during the task.

  14. Comparison of laboratory and in situ evaluation of environmental TL dosimeters

    International Nuclear Information System (INIS)

    Deme, S.; Apathy, I.; Feher, I.; Osvay, M.

    1996-01-01

    The passive environmental gamma-radiation dosimetry is mainly based on TL (thermoluminescent) dosimetry. This method offers considerable advantages due to its high precision, low cost, wide range, etc.. At the same time its application involves uncertainty caused by the dose collected during the transport from the point of annealing to the place of exposure and back to the place of evaluation. Should an accident occur read-out is delayed due to the need to transport to a laboratory equipped with a TLD reader. A portable reader capable of reading out the TL dosimeter at the place of exposure (in situ TLD reader) eliminates the disadvantages mentioned above. A microprocessor based portable TLD reader was developed by us for monitoring environmental gamma-radiation doses. Using a portable reader for in situ evaluation there are several disadvantages as well. The method requires the transport of the reader instead of dosimeters. The portable reader should be battery operated with low power consumption. Due to this requirement the temperature stabilization of the reader requests different solution as in laboratory type devices. Comparison of recently developed in situ and traditional laboratory evaluation methods of environmental TL dosimeters is given in recent paper. The comparison was made in the same conditions. The most characteristic - for environmental monitoring - numerical TL data (dose range, reproducibility, fading, self dose etc.) are given for manufactured by us CaSO 4 :Dy bulbs (portable reader) and very advantageous, high sensitive Al 2 O 3 :C dosimeters (laboratory evaluation). (author)

  15. Fast-neutron solid-state dosimeter

    International Nuclear Information System (INIS)

    Kecker, K.H.; Haywood, F.F.; Perdue, P.T.; Thorngate, J.H.

    1975-01-01

    This patent relates to an improved fast-neutron solid-state dosimeter that does not require separation of materials before it can be read out, that utilizes materials that do not melt or otherwise degrade at about 300 0 C readout temperature, that provides a more efficient dosimeter, and that can be reused. The dosimeters are fabricated by intimately mixing a TL material, such as CaSO 4 :Dy, with a powdered polyphenyl, such as p-sexiphenyl, and hot-pressing the mixture to form pellets, followed by out-gassing in a vacuum furnace at 150 0 C prior to first use dosimeters

  16. Evaluation of discrepancies between thermoluminescent dosimeter and direct-reading dosimeter results

    International Nuclear Information System (INIS)

    Shaw, K.R.

    1993-07-01

    Currently at Oak Ridge National Laboratory (ORNL), the responses of thermoluminescent dosimeters (TLDs) and direct-reading dosimeters (DRDs) are not officially compared or the discrepancies investigated. However, both may soon be required due to the new US Department of Energy (DOE) Radiological Control Manual. In the past, unofficial comparisons of the two dosimeters have led to discrepancies of up to 200%. This work was conducted to determine the reasons behind such discrepancies. For tests conducted with the TLDs, the reported dose was most often lower than the delivered dose, while DRDs most often responded higher than the delivered dose. Trends were identified in personnel DRD readings, and ft was concluded that more training and more control of the DRDs could improve their response. TLD responses have already begun to be improved; a new background subtraction method was implemented in April 1993, and a new dose algorithm is being considered. It was concluded that the DOE Radiological Control Manual requirements are reasonable for identifying discrepancies between dosimeter types, and more stringent administrative limits might even be considered

  17. The importance of using the dosimeter in medical professionals in the hemodynamic service

    International Nuclear Information System (INIS)

    Melo, Francisca A. de; Victor Filho, Edgard; Silva, Carla V. da; Santos, Tayline T. dos; Guerra, Decio C.

    2014-01-01

    The objective of this study is to evaluate the medical exposure to ionizing radiation of X type in a interventional radiology service, of an university hospital, making a correlation with the importance of using dosimeters for monitoring the effective dose in individuals occupationally exposed (IOE). It was performed an analysis of radiation doses in two stages: the first there was not guidance on the need of using dosimeters; in the second time the professionals performed all procedures carrying the dosimeter. The result showed an average effective dose of professionals / year of 8.60 mSv at first moment, against a dose of 27.41 mSv in the second time after the routine of use the dosimeters, surpassing, in this second phase, the annual dose rate allowed by current radiation protection legislation, which calls for 20 mSv / year for professional. The comparison result in an increase of effective dose of professionals in nearly 300%. It is concluded that the implementation a continuing education project, including awareness of the importance in daily use dosimeter, shows up as a solution for optimizing the dose of these occupationally exposed individuals

  18. Thermoluminescent dosimeter system

    International Nuclear Information System (INIS)

    Felice, P.E.; Gonzalez, J.L.; Seidel, J.G.

    1979-01-01

    An improved thermoluminescent dosimeter system and apparatus for sensing alpha particle emission is described. A thermoluminescent body is sealed between a pair of metallized plastic films. The dosimeter is mounted within a protective inverted cup or a tube closed at one end, which is disposed in a test hole for exposure to radioactive radon gas which is indicaive of uranium deposits

  19. A design of ambient dose equivalent dosimeter and its dosimetric performance

    International Nuclear Information System (INIS)

    Zhao Shian; Ou Xiangming; Li Kaibao

    1997-01-01

    Objective: To design an ambient dose equivalent dosimeter with digital display for radiation protection, which is based on the definition of the new operational radiation quantity for environmental monitoring-ambient dose equivalent recommended by the International Commission on Radiation Units and Measurements (ICRU) Report 39. Methods: Considering the energy response of the instrument, the inner wall of ionizing chamber is coated with gum graphite added with a bit of metal powder. Results: Using this chamber, measurement of H * (10) for photon radiation with unknown spectrum distribution is possible in the energy range from 47 keV to 230 keV with an uncertainty of better than 5%. The configuration, technology and dosimetric performance of the chamber and automatic functions of the reader are presented. Conclusion: The ambient dose equivalent dosimeter can be used as not only a working reference dosimeter, but also a field dosimeter for radiation protection because the readings are expressed directly in ambient dose equivalent and averaged automatically in the period of measurement. Also, its power is supplied by battery for the portable purpose and the readings are displayed on the screen with light-background for dim field

  20. Flame-sintered ceramic exoelectron dosimeter samples

    International Nuclear Information System (INIS)

    Petel, M.; Holzapfel, G.

    1979-01-01

    New techniques for the preparation of integrating solid state dosimeters, particularly exoelectron dosimeters, have been initiated. The procedure consists in melting the powdered dosimeter materials in a hot, fast gas stream and depositing the ceramic layer. The gas stream is generated either through a chemical flame or by an electrical arc plasma. Results will be reported on the system Al 2 O 3 /stainless steel as a first step to a usable exoelectron dosimeter

  1. Conceptual design of the SMART dosimeter

    Science.gov (United States)

    Johnson, Erik B.; Vogel, Sam; Frank, Rebecca; Stoddard, Graham; Vera, Alonzo; Alexander, David; Christian, James

    2017-08-01

    Active dosimeters for astronauts and space weather monitors are critical tools for mitigating radiation induced health issues or system failure on capital equipment. Commercial spaceflight, deep space flight, and satellites require smarter, smaller, and lower power dosimeters. There are a number of instruments with flight heritage, yet as identified in NASA's roadmaps, these technologies do not lend themselves to a viable solution for active dosimetry for an astronaut, particularly for deep space missions. For future missions, nano- and micro-satellites will require compact instruments that will accurately assess the radiation hazard without consuming major resources on the spacecraft. RMD has developed the methods for growing an advanced scintillation material called phenylcarbazole, which provides pulse shape discrimination between protons and electrons. When used in combination with an anti-coincidence detector system, an assessment of the dose from charged ions and neutral particles can be determined. This is valuable as damage on a system (such as silicon or tissue) is dependent on the particle species. Using this crystal with readout electronics developed in partnership with COSMIAC at the University of New Mexico, the design of the Small Mixed field Autonomous Radiation Tracker (SMART) Dosimeter consists of a low-power analog to digital conversion scheme with low-power digital signal processing algorithms, which are to be implemented within a compact system on a chip, such as the Xilinx Zynq series. A review of the conceptual design is presented.

  2. Passive dosimeters other than film and TLDs [thermoluminescent dosimeter

    International Nuclear Information System (INIS)

    Hankins, D.E.

    1986-01-01

    This presentation will describe CR-39 plastic as a personnel neutron dosimeter. Recent research at LLNL and elsewhere has resulted in the development of a dosimetry system that is superior to any personnel neutron dosimeter previously available. The author describes the features of the dosimetry system and the new etching procedures and techniques in detail. Most of the research was done at the LLNL and has been supported as a part of the DOE Neutron Dosimetry Upgrade Program. 10 refs., 4 figs., 1 tab

  3. Neutron dosimeter

    International Nuclear Information System (INIS)

    Bartko, J.; Schoch, K.F. Jr.; Congedo, T.V.; Anderson, S.L. Jr.

    1989-01-01

    This patent describes a nuclear reactor. It comprises a reactor core; a thermal shield surrounding the reactor core; a pressure vessel surrounding the thermal shield; a neutron dosimeter positioned outside of the thermal shield, the neutron dosimeter comprising a layer of fissile material and a second layer made of a material having an electrical conductivity which permanently varies as a function of its cumulative ion radiation dose; and means, outside the pressure vessel and electrically connected to the layer of second material, for measuring electrical conductivity of the layer of second material

  4. Optical CT imaging of solid radiochromic dosimeters in mismatched refractive index solutions using a scanning laser and large area detector.

    Science.gov (United States)

    Dekker, Kurtis H; Battista, Jerry J; Jordan, Kevin J

    2016-08-01

    The practical use of the PRESAGE® solid plastic dosimeter is limited by the inconvenience of immersing it in high-viscosity oils to achieve refractive index matching for optical computed tomography (CT) scanning. The oils are slow to mix and difficult to clean from surfaces, and the dosimeter rotation can generate dynamic Schlieren inhomogeneity patterns in the reference liquid, limiting the rotational and overall scan speed. Therefore, it would be beneficial if lower-viscosity, water-based solutions with slightly unmatched refractive index could be used instead. The purpose of this work is to demonstrate the feasibility of allowing mismatched conditions when using a scanning laser system with a large acceptance angle detector. A fiducial-based ray path measurement technique is combined with an iterative CT reconstruction algorithm to reconstruct images. A water based surrounding liquid with a low viscosity was selected for imaging PRESAGE® solid dosimeters. Liquid selection was optimized to achieve as high a refractive index as possible while avoiding rotation-induced Schlieren effects. This led to a refractive index mismatch of 6% between liquid and dosimeters. Optical CT scans were performed with a fan-beam scanning-laser optical CT system with a large area detector to capture most of the refracted rays. A fiducial marker placed on the wall of a cylindrical sample occludes a given light ray twice. With knowledge of the rotation angle and the radius of the cylindrical object, the actual internal path of each ray through the dosimeter can be calculated. Scans were performed with 1024 projections of 512 data samples each, and rays were rebinned to form 512 parallel-beam projections. Reconstructions were performed on a 512 × 512 grid using 100 iterations of the SIRT iterative CT algorithm. Proof of concept was demonstrated with a uniformly attenuating solution phantom. PRESAGE® dosimeters (11 cm diameter) were irradiated with Cobalt-60 irradiator to achieve

  5. Optical CT imaging of solid radiochromic dosimeters in mismatched refractive index solutions using a scanning laser and large area detector

    Energy Technology Data Exchange (ETDEWEB)

    Dekker, Kurtis H., E-mail: kdekker2@uwo.ca [Department of Medical Biophysics, Schulich School of Medicine and Dentistry, The University of Western Ontario, London, Ontario N6A 5C1 (Canada); Battista, Jerry J.; Jordan, Kevin J. [Departments of Medical Biophysics and Oncology, Schulich School of Medicine and Dentistry, The University of Western Ontario, London, Ontario N6A 5C1, Canada and Department of Physics and Engineering, London Regional Cancer Program, London Health Sciences Centre, 790 Commissioners Road East, London, Ontario N6A 4L6 (Canada)

    2016-08-15

    Purpose: The practical use of the PRESAGE® solid plastic dosimeter is limited by the inconvenience of immersing it in high-viscosity oils to achieve refractive index matching for optical computed tomography (CT) scanning. The oils are slow to mix and difficult to clean from surfaces, and the dosimeter rotation can generate dynamic Schlieren inhomogeneity patterns in the reference liquid, limiting the rotational and overall scan speed. Therefore, it would be beneficial if lower-viscosity, water-based solutions with slightly unmatched refractive index could be used instead. The purpose of this work is to demonstrate the feasibility of allowing mismatched conditions when using a scanning laser system with a large acceptance angle detector. A fiducial-based ray path measurement technique is combined with an iterative CT reconstruction algorithm to reconstruct images. Methods: A water based surrounding liquid with a low viscosity was selected for imaging PRESAGE® solid dosimeters. Liquid selection was optimized to achieve as high a refractive index as possible while avoiding rotation-induced Schlieren effects. This led to a refractive index mismatch of 6% between liquid and dosimeters. Optical CT scans were performed with a fan-beam scanning-laser optical CT system with a large area detector to capture most of the refracted rays. A fiducial marker placed on the wall of a cylindrical sample occludes a given light ray twice. With knowledge of the rotation angle and the radius of the cylindrical object, the actual internal path of each ray through the dosimeter can be calculated. Scans were performed with 1024 projections of 512 data samples each, and rays were rebinned to form 512 parallel-beam projections. Reconstructions were performed on a 512 × 512 grid using 100 iterations of the SIRT iterative CT algorithm. Proof of concept was demonstrated with a uniformly attenuating solution phantom. PRESAGE® dosimeters (11 cm diameter) were irradiated with Cobalt-60

  6. Radiation detectors for personnel monitoring - current developments and future trends

    International Nuclear Information System (INIS)

    Kannan, S.

    2003-01-01

    The radiation detectors for personnel monitoring range from the conventional passive dosimeters like the film badge and the TLD, to sophisticated active dosimeters for integrated gamma, beta and neutron dose measurement. With the availability of high accuracy active dosimeters, the process of personnel monitoring, acceptability among radiation workers, record keeping and dose control have become more simplified. However the high level of sophistication in the active dosimeter has its own inevitable price tag and the new breed of active dosimeters are prohibitively costly. The silver lining, in the otherwise dark cost scenario of these dosimeters is the potential for cost reduction at least in some of the dosimeters in the near future

  7. Development of web monitoring radiation area monitor

    International Nuclear Information System (INIS)

    Jung, Hoon Jin; Lee, Jun Hee; Namkoong, Phil; Lee, Dong Hoon; Lee, Su Hong; Lee, Gun Bae

    2005-01-01

    Recently the increasing number of radioisotope industry and nuclear facility have ever raised the possibility of radiation safety accident. As such a result, radioisotope companies and nuclear facility operators have become to be much interested in radiation area monitoring for efficient radiation protection. At present, almost of the radiation area monitors which are imported products are outdated in aspect of their functions. Diversification of the monitoring work is urgently demanding additional functions to be added. Thus we have developed new-type digital area monitor which enables remote web monitoring with image and radiation dose rate value at distant places through using internet, the latest IT technology, and radiation measurement technology

  8. Improved sample holders for the PMMA dosimeters

    International Nuclear Information System (INIS)

    Kobayashi, Toshikazu; Sone, Koji; Iso, Katsuaki

    1994-01-01

    PMMA dosimeters are widely used for high dose dosimetry. Dose is determined by measuring the change in optical density of the irradiated PMMA dosimeter element. Measurement precision depends on the mounting method of a dosimeter element in the sample room of a spectrophotometer. We tried to prepare three types of holders, (holders A, B and C in Figs. 1-3), according to the shape of PMMA dosimeter elements. We measured optical density of the irradiated PMMA dosimeter elements by using the three types of holders. It is revealed that the holder of the type A gives more precise results for the Red 4034 or Gammachrome YR dosimeter than that of the type B. The measurements with a spectrophotometer using the type C holder gives better results for the Red acrylic dosimeter than the case of the measurements by the exclusive reader. (author)

  9. The intelligence of dosimeter for ionization radiation

    International Nuclear Information System (INIS)

    He Jinglun

    1992-01-01

    The connection of dosimeter with microcomputer system is described, which has the functions of sampling, data handling, display and printing dose values in legal units of measurement. The accuracy and speed of measurement for dosimeters are also raised, thereby the dosimeters are made to have intelligence and the application range of dosimeter is enlarged

  10. Feasibility study of a photoconductor based dosimeter for quality assurance in radiotherapy

    Science.gov (United States)

    Lee, Y. K.; Kim, S. W.; Kim, J. N.; Kang, Y. N.; Kim, J. Y.; Lee, D. S.; Kim, K. T.; Han, M. J.; Ahn, K. J.; Park, S. K.

    2017-09-01

    With the recent market entries of new types of linear accelerators (LINACs) with a multi leaf collimator (MLC) mounted on them, high-precision radiosurgery applying a LINAC to measure high-dose radiation on the target region has been gaining popularity. Systematic and accurate quality assurance (QA) is of vital important for high-precision radiosurgery because of its increased risk of side effects including life-threatening ones such as overexposure of healthy tissues to high-dose radiation beams concentrated on small areas. Therefore, accurate dose and dose-distribution measurements are crucial in the treatment procedure. The accurate measurement of the properties of beams concentrated on small areas requires high-precision dosimeters capable of high-resolution output and dose mapping as well as accurate dosimetry in penumbra regions. In general, the properties of beams concentrated on small areas are measured using thermos luminescent dosimeters (TLD), diode detectors, ion chambers, diamond detectors, or films, and many papers have presented the advantages and disadvantages of each of these detectors for dosimetry. In this study, a solid-state photoconductor dosimeter was developed, and its clinical usability was tested by comparing its relative dosimetric performance with that of a conventional ion chamber. As materials best-suited for radiation dosimeters, four candidates namely lead (II) iodide (PbI2), lead (II) oxide (PbO), mercury (II) iodide (HgI2), and HgI2/ titanium dioxide (TiO2) composite, the performances of which were proved in previous studies, were used. The electrical properties of each candidate material were examined using the sedimentation method, one of the particle-in-binder (PIB) methods, and unit-cell-type prototypes were fabricated. The unit-cell samples thus prepared were cut into specimens of area 1 × 1 cm2 with 400-μ m thickness. The electrical properties of each sample, such as sensitivity, dark current, output current, rising time

  11. Acoustic evaluation of polymer gel dosimeters

    International Nuclear Information System (INIS)

    Mather, M.L.; De Deene, Y.; Baldock, C.; Whittaker, A.K.

    2002-01-01

    Advances in radiotherapy treatment techniques such as intensity modulated radiotherapy are placing increasing demands on radiation dosimetry for verification of dose distributions in 3D. In response, polymer gel dosimeters that are capable of recording dose distributions in 3D are currently being developed. Recently, a new technique for evaluation of absorbed dose distributions in these dosimeters using ultrasound was introduced. The current work aims to demonstrate the potential of ultrasound as an evaluation technique for polymer gel dosimeters and to investigate the ultrasound properties of two different dosimeter formulations, PAG and MAGIC gels

  12. Water-resistant alanine-EPR dosimeter alanpol

    International Nuclear Information System (INIS)

    Peimel-Stuglik, Zofia; Bryl-Sandelewska, Teresa; Mirkowski, Krzysztof; Sartowska, Bozena

    2009-01-01

    Alanpol-water-resistant alanine-electron paramagnetic resonance (EPR) dosimeter consisted of cheap DL-α-alanine (9.8-27%) suspended in polyethylene matrix was presented. The rods (O=2.8 mm) were extruded from a hot mixture of alanine and low-density polyethylene. No grinding or crushing was used for alanine preparation. An orientation of cylindrical crystals, up to 300 μm long in parallel to the rod axis was responsible for some differences in a shape of EPR signal. These differences had no negative consequences for dosimetric applications. Signal-to-dose dependence was linear up to 10 kGy. Standard deviation of dosimetric answer was up to ±1.8% and up to 2.4% for dosimeters with 9.8% and 27% of DL-α-alanine, respectively. Irradiation temperature coefficient for both dosimeters was equal 0.2%/ deg. C. Hydrophobic properties of polyethylene and small number of alanine crystals located on the surface of the rod led to high resistance of dosimeters to water and humidity. The 24 h soaking of irradiated dosimeters in liquid water-reduced EPR signals by 3-4% and by 2-3% for dosimeters with 27% and 9.8% of DL-α-alanine, respectively. Three month storage time of irradiated dosimeters in room conditions decreases EPR signal for ∼3%.

  13. Nuclear radiation monitoring instruments for personnel in nuclear disaster for defence needs

    International Nuclear Information System (INIS)

    Bhatnagar, P.K.; Vaijapurkar, S.G.; Yadav, Ashok

    2005-01-01

    Ever since the tragedy of nuclear device exploding over Japan by USA in 1945 awareness exists amongst the armed forces personnel all over the world that a requirement of implementing radiological protection is imminent. Towards this adoption of radiological safety programme is a criterion. In a nuclear war disaster scenario, one encounters initial nuclear radiation (gamma and neutron radiations), gamma radiations from fallout, heat and blast. At certain distances Tanks/ armoured vehicles will survive and 4 R/s radiation level sensing to actuate relays for closing the ports of vehicles is essential, leading to reduction in inhalation, ingestion of fallout radioactivity and reduction in radiation dose received by occupants of the vehicle. Towards this sturdy radiation monitors to indicate gamma dose rate of the order of 1000 R/h, gamma and neutron dosimeters of the order of 1000 cGy with reading instruments are to be developed. These must work in harsh environment and sustain JSS 55555 conditions of army. Defence Laboratory, Jodhpur over past one decade has been involved in developing personnel, area and field monitoring instruments like dosimeters, survey meters, which are useful, acceptable to army personnel, armoured and personnel carrier vehicles, field structures/shelters. Technology transfer after satisfaction of armed forces, product ionisation and supply, maintenance, training has been the endeavor of the DRDO. Herein it is proposed to highlight the techno electronics nuclear radiation monitoring sensors and associated electronics systems developed first time in the country and productionised in bulk for Services for implementing personnel protection. The sensors developed and described are - Radiophotoluminescent Glass (RPLG) for gamma radiation dosimetry , neutron sensitive PIN diode for fast neutron dosimetry, gamma radiation sensitive PIN diode, superheated liquid neutron and gamma sensors. The dosimeter, dose rate meter and field/area instruments are

  14. SDI-100 radiation dosimeter

    International Nuclear Information System (INIS)

    Zheng Zheng; Zhao Yongfu; Dai Honggui

    1995-01-01

    An intelligent radiation dosimeter, with such functions as signal collection and data processing, store, print and display, has been developed. Its detector is made of a micro-semiconductor. This dosimeter can be used in laboratories for agricultural 60 Co irradiators, radiotherapeutic facilities and other small and medium-size 60 Co irradiators

  15. Evaluation of gamma radiation cumulative exposure monitoring system in the Nuclear Technology Development Center (CDTN) with thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Domingues, Danieli S.; Barreto, Alberto A.; Siqueira, Gessilane M.; Oliveira, Eugênio M.; Silva, Robson L.; Belo, Luiz Cláudio M.

    2017-01-01

    Cumulative exposure to gamma radiation is monitored at the Nuclear Technology Development Center - CDTN, located on the campus of the Federal University of Minas Gerais in Belo Horizonte - MG, since 1986. This monitoring is part of the Institution's Environmental Monitoring Program. The measurement is carried out through the use of thermoluminescent dosimeters (TLD's), which are placed in the perimeter of the CDTN property, distanced approximately 50 m each other, and at the air sampling stations, totaling 36 sampling points. The TLDs are positioned one meter from the ground, inside a PVC pipe where they remain exposed for a period of approximately three months. The objective of this work is to present and evaluate the results of this monitoring by investigating the dose values of gamma radiation (Dose Equivalent to Photons) obtained. Another important issue presented is about some failures identified in certain periods. There were problems with the sampling methodology and changes were proposed in the shelters to avoid contact of TLDs with excess moisture, which can compromise the results and the equipment of measurement of the thermoluminescence of crystals. The change was made at a sampling point and the results were compared with the historical data in the last 30 years. (author)

  16. Evaluation of gamma radiation cumulative exposure monitoring system in the Nuclear Technology Development Center (CDTN) with thermoluminescent dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Domingues, Danieli S.; Barreto, Alberto A.; Siqueira, Gessilane M.; Oliveira, Eugênio M.; Silva, Robson L.; Belo, Luiz Cláudio M., E-mail: danielisilva@ymail.com, E-mail: aab@cdtn.br, E-mail: gessilane.siqueira@cdtn.br, E-mail: emo@cdtn.br, E-mail: rls@cdtn.br, E-mail: lcmb@cdtn.br [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)

    2017-07-01

    Cumulative exposure to gamma radiation is monitored at the Nuclear Technology Development Center - CDTN, located on the campus of the Federal University of Minas Gerais in Belo Horizonte - MG, since 1986. This monitoring is part of the Institution's Environmental Monitoring Program. The measurement is carried out through the use of thermoluminescent dosimeters (TLD's), which are placed in the perimeter of the CDTN property, distanced approximately 50 m each other, and at the air sampling stations, totaling 36 sampling points. The TLDs are positioned one meter from the ground, inside a PVC pipe where they remain exposed for a period of approximately three months. The objective of this work is to present and evaluate the results of this monitoring by investigating the dose values of gamma radiation (Dose Equivalent to Photons) obtained. Another important issue presented is about some failures identified in certain periods. There were problems with the sampling methodology and changes were proposed in the shelters to avoid contact of TLDs with excess moisture, which can compromise the results and the equipment of measurement of the thermoluminescence of crystals. The change was made at a sampling point and the results were compared with the historical data in the last 30 years. (author)

  17. Study of a plastic detector as a neutron dosimeter

    International Nuclear Information System (INIS)

    Cuauhtecatl Hernandes, V.

    1982-01-01

    Studies were carried out through nuclear reactions (n, p); (n, heavy ion), and (n,α) of the dosimetric properties of CR 39 commercial brand polymers. A system was devised for calculating the number of neutron induced nuclear reactions and geometric absolute efficiency factors. Feasibility of the utilization of CR 39 in monitoring and in personnel dosimeters is also discussed. (author)

  18. Hanford beta-gamma personnel dosimeter prototypes and evaluation

    International Nuclear Information System (INIS)

    Fix, J.J.; Holbrook, K.L.; Soldat, K.L.

    1983-04-01

    Upgraded and modified Hanford dosimeter prototypes were evaluated for possible use at Hanford as a primary beta-gamma dosimeter. All prototypes were compatible with the current dosimeter card and holder design, as well as processing with the automated Hanford readers. Shallow- and deep-dose response was determined for selected prototypes using several beta sources, K-fluorescent x rays and filtered x-ray techniques. All prototypes included a neutron sensitive chip. A progressive evaluation of the performance of each of the upgrades to the current dosimeter is described. In general, the performance of the current dosimeter can be upgraded using individual chip sensitivity factors to improve precision and an improved algorithm to minimize bias. The performance of this dosimeter would be adequate to pass all categories of the ANSI N13.11 performance criteria for dosimeter procesors, provided calibration techniques compatible with irradiations adopted in the standard were conducted. The existing neutron capability of the dosimeter could be retained. Better dosimeter performance to beta-gamma radiation can be achieved by modifying the Hanford dosimeter so that four of the five chip positions are devoted to calculating these doses instead of the currently used two chip positions. A neutron sensitive chip was used in the 5th chip position, but all modified dosimeter prototypes would be incapable of discriminating between thermal and epithermal neutrons. An improved low energy beta response can be achieved for the current dosimeter and all prototypes considered by eliminating the security credential. Further improvement can be obtained by incorporating the 15-mil thick TLD-700 chips

  19. The neutron response of a 7 LiF thermoluminescent dosimeter incorporated in the UKAEA criticality dosimeter

    International Nuclear Information System (INIS)

    Eid, A.M.; Delafield, H.J.

    1976-04-01

    There are practical advantages in incorporating a 7 LiF thermoluminescent dosimeter (TLD) for the measurement of γ-ray dose, into the personnel criticality dosimeter. This paper investigated the corrections necessary for the inherent direct response of the TLD neutrons, and its enhanced indirect response from prompt γ-rays resulting from neutron interactions with the metallic foils contained in the criticality dosimeter. The response of the TLD to fast fission neutrons was measured to be 0.02 γ rad/n rad. The indirect response of the TLD to thermal neutrons was measured to be 4.8 x 10 -10 rad n -1 cm 2 for dosimeters exposed in free air, and 7 x 10 -10 rad n -1 cm 2 for dosimeters worn on the body respectively. Application of these correction factors to TLD measurements made at International Dosimetry Intercomparisons (sponsored by the I.A.E.A.) gave improved agreement with the values given by other participants. (author)

  20. External gamma radiation monitoring in the environs of Kaiga Generating Station (KGS), using thermoluminescent dosimeters, during the period 1989-2003

    International Nuclear Information System (INIS)

    Basu, A.S.; Chougaonkar, M.P.; Mayya, Y.S.; Puranik, V.D.; Reji, T.K.; Ravi, P.M.; Hegde, A.G.

    2005-05-01

    This publication reports the results of external gamma radiation monitoring using Thermoluminescent Dosimeters (TLDs), in the environs of Kaiga Generating Station (KGS) during its preoperational survey between October 1989 and June 1998. The report also presents quarterly and annual values of air dose during the operational phase of the station between July 1998 and Dec. 2003 around the environmcnt of KGS. The results of TLD analysis, during the period October 1989-June 1998, indicate that the average annual air dose for the locations monitored, was 502± 91 μGy/a. The general background of the environs around Kaiga during the operational period, i.e. July 1998 and Dec. 2003, between 2.3 km. to 32km. has been found to be 509±74 μGy/a. The report discusses the methodology and different analyses carried out. (author)

  1. Silver nitrate based gel dosimeter

    International Nuclear Information System (INIS)

    Titus, D; Samuel, E J J; Srinivasan, K; Roopan, S M; Madhu, C S

    2017-01-01

    A new radiochromic gel dosimeter based on silver nitrate and a normoxic gel dosimeter was investigated using UV-Visible spectrophotometry in the clinical dose range. Gamma radiation induced the synthesis of silver nanoparticles in the gel and is confirmed from the UV-Visible spectrum which shows an absorbance peak at around 450 nm. The dose response function of the dosimeter is found to be linear upto12Gy. In addition, the gel samples were found to be stable which were kept under refrigeration. (paper)

  2. Personnel neutron dosimeter for use in a plutonium processing plant

    International Nuclear Information System (INIS)

    Brunskill, R.T.; Hwang, F.S.W.

    1978-01-01

    A thermoluminesence dosimeter for personnel neutron dose measurement, which is based on the albedo principle, has been developed at Windscale works. The dosimeter has been calibrated against a 238 Pu/Be neutron source using different degrees of moderation and against a variety of neutron spectra prevailing in different areas of the Plutonium Finishing Plant. The dosimeter consists of two identical parts in which the sensitive elements are graphite discs which have thermoluminescent crystals sealed to the plane faces with a high temperature resin. The graphite discs are supported in teflon washers which fit into a body of tufnol. A circular insert of boronated polythene in each tufnol body provides a thermal neutron absorber for the sensitive element in the other half of the dosimeter. Natural lithium borate was used as the neutron sensitive phosphor and a lithium borate made from isotopes 7 Li (99.9%) and 11 B (99.2%) as the neutron insensitive materials. Neutron-sensitive lithium borate is sealed to one face of each disc and the neutron-insensitive material to the opposite face. The dosimeter is so assembled that the neutron-sensitive faces both lie in the central plane. The design is such that one neutron sensitive face responds to the incident flux of neutron only while the other responds to the albedo flux

  3. Feasibility of smartphone diaries and personal dosimeters to quantitatively study exposure to ultraviolet radiation in a small national sample

    DEFF Research Database (Denmark)

    Køster, Brian; Søndergaard, Jens; Nielsen, Jesper B

    2015-01-01

    studies and of smartphones as a data collection tool. MATERIALS AND METHODS: Participants were sent a dosimeter which they wore for 7 days, received a short diary questionnaire by text message each day and subsequently a longer questionnaire. Correlation between responses from questionnaire, smartphone...... diaries and dosimeters were examined. RESULTS: This study shows a 99.5% return rate (n = 205) of the dosimeters by ordinary mail and high response-rates for a smartphone questionnaire dairy. Correlation coefficients for outdoor-time reported through smartphones and dosimeters as average by week 0.62 (0...... that dosimeter studies are feasible in national settings and that smartphones are a useful tool for monitoring and collecting UV behavior data. CONCLUSION: We found diary data reported on a daily basis through smartphones more strongly associated with actual outdoor time than questionnaire data. Our results...

  4. Redox-Phen solution: A water equivalent dosimeter for UVA, UVB and X-rays radiation

    Science.gov (United States)

    Marini, A.; Ciribolla, C.; Lazzeri, L.; d'Errico, F.

    2018-06-01

    Polysulphone films are the only type of UV passive dosimeters that are widely adopted for research and personal monitoring. Even though many studies concentrated on the development and characterization of these films, they still present some shortcomings. The more important limitations of them are that they can measure only UVB radiations and that they change color at 330 nm, requiring special equipment to read them. To overcome these limitations we developed an aqueous dosimeter that is sensitive to UVA, UVB and X-rays named Redox-Phen solution. This dosimeter is inexpensive and water equivalent, being made of more than 99 wt% of water. It changes color in the visible region upon irradiation, thus it can be measured via simple optical method, and an evaluation of the exposition can be made also by naked eyes.

  5. The development of wireless radiation dose monitoring using smart phone

    Energy Technology Data Exchange (ETDEWEB)

    Lee, Jin Woo; Jeong, Gyo Seong; Lee, Yun Jong [Advanced Radiation Technology Institute, Korea Atomic Energy Research Institute, Jeongeup (Korea, Republic of); Kim, Chong Yeal [Chonbuk National University, Jeonju (Korea, Republic of); Lim, Chai Wan [REMTECH, Seoul (Korea, Republic of)

    2016-11-15

    Radiation workers at a nuclear facility or radiation working area should hold personal dosimeters. some types of dosimeters have functions to generate audible or visible alarms to radiation workers. However, such devices used in radiation fields these days have no functions to communicate with other equipment or the responsible personnel. our project aims at the development of a remote wireless radiation dose monitoring system that can be utilized to monitor the radiation dose for radiation workers and to notify the radiation protection manager of the dose information in real time. We use a commercial survey meter for personal radiation measurement and a smart phone for a mobile wireless communication tool and a Beacon for position detection of radiation workers using Blue tooth communication. In this report, the developed wireless dose monitoring of cellular phone is introduced.

  6. The task of official personal monitoring in Germany using electronic dosimetry systems

    International Nuclear Information System (INIS)

    Huebner, Stephan; Wahl, Wolfgang; Busch, Frank; Martini, Ekkehard

    2008-01-01

    Full text: Since the establishment of the first German personal monitoring services as competent measuring bodies in the year 1952, official personal dosimetry is carried out using passive dosimeters such as film batches, RPL- and TL-dosimeters solely. On the other hand, electronic dosimeters are in use in some big institutions like Nuclear Power Plants, hospitals or industrial units for operational purposes. In most cases, these dosimeters are regulated by competent authorities. For more than 20 years electronic dosimeters proved their worth of being appropriate personal dosimeters. Since 2001 concepts to implement electronic personal dosimeters into the official individual monitoring of occupational exposed workers were developed in different research projects. The EU market of personal dosimetry changes to an open and competitive one, the number of outside workers, especially during the outages of Nuclear Power Plants increases, the landscape of customers is getting more and more heterogeneous. Being able to face these tasks of a sustainable personal monitoring requires the introduction of modern electronic dosimeters into to the official monitoring. Doing so, the needed prompt exchange of dose-data between different monitoring services as well as between the customers and the related monitoring service can be warranted. In cooperation with the industry, competent authorities and a research centre a method for official dosimetry using electronic dosimetry systems was developed, realised and tested successfully by the three big monitoring services of Germany. These investigations are supported by the German Federal Ministry for the Environment, Nature Conservation and Nuclear Safety. For this purpose a network between customers and monitoring services was built up in order to monitor people, who work in different places related to different measuring bodies in only one period of surveillance. (author)

  7. New Neutron Dosimeter

    CERN Multimedia

    2001-01-01

    CERN has been operating an Individual Dosimetry Service for neutrons for about 35 years. The service was based on nuclear emulsions in the form of film packages which were developed and scanned in the Service. In 1999, the supplier of theses packages informed CERN that they will discontinue production of this material. TIS-RP decided to look for an external service provider for individual neutron dosimetry. After an extensive market survey and an invitation for tender, a supplier that met the stringent technical requirements set up by CERN's host states for personal dosimeters was identified. The new dosimeter is based on a track-etching technique. Neutrons have the capability of damaging plastic material. The microscopic damage centres are revealed by etching them in a strong acid. The resulting etch pits can be automatically counted and their density is proportional to dose equivalent from neutrons. On the technical side, the new dosimeter provides an improved independence of its response from energy and th...

  8. Online monitoring of absorbed dose in undulator magnets with RADFET dosimeters at FERMI@Elettra

    International Nuclear Information System (INIS)

    Fröhlich, L.; Casarin, K.; Quai, E.; Holmes-Siedle, A.; Severgnini, M.; Vidimari, R.

    2013-01-01

    The FERMI@Elettra free-electron laser, based on a 1.3 GeV electron linac, requires the monitoring of radiation doses up to a few kGy for the protection of sensitive equipment such as permanent magnet undulators. A new dosimetry system DOSFET-L01, employing an array of RADFETs spread throughout the accelerator, was developed. So far, the system has performed flawlessly for almost two years, taking one dose reading per minute around the clock. The REM RFT-300 sensors were set in zero-bias mode, i.e. with all electrodes grounded during exposure. This choice of mode allows the measurement of a high range of integrated doses – up to a few kGy. The paper describes the new read-out system and its application, calibration measurements in cobalt-60 and 6 MeV bremsstrahlung radiation sources giving rise to a novel response function, and new data on “fade” under the zero-bias mode of use for over 300 days at room temperature. Regular readings from 28 RADFETs placed within seven undulators over the first 20 months of operation of the accelerator demonstrate how the system tracks and locates periods of high and low dose rate and thereby contributes to the protection from beam loss. The readings from the RADFET system are found to be in good agreement with Gafchromic EBT2 film dosimeters. Based on the results reported, the choice of bias mode may be revised so as to reduce fade and improve the accuracy conferred by a positive-bias mode. -- Highlights: ► We developed a system for online dosimetry with RADFET sensors under zero bias. ► The system is calibrated for doses up to 10 kGy with REM RFT-300 sensors. ► We collected data on fade for over 300 days from irradiation. ► We present undulator dose measurements for 20 months of operation of FERMI@Elettra. ► Dose measurements are in good agreement with chemical film dosimeters

  9. Performance testing of extremity dosimeters, Study 2

    International Nuclear Information System (INIS)

    Harty, R.; Reece, W.D.; Hooker, C.D.

    1990-04-01

    The Health Physics Society Standards Committee (HPSSC) Working Group on Performance Testing of Extremity Dosimeters has issued a draft of a proposed standard for extremity dosimeters. The draft standard proposes methods to be used for testing dosimetry systems that determine occupational radiation dose to the extremities and the performance criterion used to determine compliance with the standard. Pacific Northwest Laboratory (PNL) has conducted two separate evaluations of the performance of extremity dosimeter processors to determine the appropriateness of the draft standard, as well as to obtain information regarding the performance of extremity dosimeters. Based on the information obtained during the facility visits and the results obtained from the performance testing, it was recommended that changes be made to ensure that the draft standard is appropriate for extremity dosimeters. The changes include: subdividing the mixture category and the beta particle category; eliminating the neutron category until appropriate flux-to-dose equivalent conversion factors are derived; and changing the tolerance level for the performance criterion to provide consistency with the performance criterion for whole body dosimeters, and to avoid making the draft standard overly difficult for processors of extremity dosimeters to pass. 20 refs., 10 figs., 6 tabs

  10. New Generation of self-calibrated SS/EPR dosimeters: Alanine/EPR dosimeters

    International Nuclear Information System (INIS)

    Yordanov, N.D.; Gancheva, V.

    1999-01-01

    A new type of solid state/EPR dosimeters is described. Principally, it contains radiation sensitive diamagnetic material, some quantity of EPR active, but radiation insensitive, substance (for example Mn 2+ /MgO) and a binding material. In the present case alanine is used as a radiation sensitive substance. With this dosimeter, the EPR spectra of alanine and Mn 2+ are simultaneously recorded and the calibration graph represents the ratio of alanine versus Mn 2+ EPR signal intensity as a function of absorbed dose. In this way the reproducibility of the results is expected to be improved significantly including their intercomparison among different laboratories. Homogeneity of the prepared dosimeters and their behaviour (fading of EPR signals with time, influence of different meteorological conditions) show satisfactory reproducibility and stability with time. Because two different EPR active samples are recorded simultaneously, the influence of some instrument setting parameters (microwave power, modulation amplitude and modulation frequency) on the ratio I alanine /I Mn is also investigated. (author)

  11. A fibre optic dosimeter customised for brachytherapy

    Energy Technology Data Exchange (ETDEWEB)

    Suchowerska, N. [Department of Radiation Oncology, Royal Prince Alfred Hospital, Camperdown, NSW 2050 (Australia); School of Physics, University of Sydney, NSW 2006 (Australia)], E-mail: Natalka@email.cs.nsw.gov.au; Lambert, J.; Nakano, T. [Department of Radiation Oncology, Royal Prince Alfred Hospital, Camperdown, NSW 2050 (Australia); School of Physics, University of Sydney, NSW 2006 (Australia); Law, S. [School of Physics, University of Sydney, NSW 2006 (Australia); Optical Fibre Technology Centre, University of Sydney, 206 National Innovation Centre, Australian Technology Park, Eveleigh, NSW 1430 (Australia); Elsey, J. [Bandwidth Foundry Pty Ltd, Australian Technology Park, NSW, 1430 (Australia); McKenzie, D.R. [School of Physics, University of Sydney, NSW 2006 (Australia)

    2007-04-15

    In-vivo dosimetry for brachytherapy cancer treatment requires a small dosimeter with a real time readout capability that can be inserted into the patient to determine the dose to critical organs. Fibre optic scintillation dosimeters, consisting of a plastic scintillator coupled to an optical fibre, are a promising dosimeter for this application. We have implemented specific design features to optimise the performance of the dosimeter for specific in-vivo dosimetry during brachytherapy. Two sizes of the BrachyFOD{sup TM} scintillation dosimeter have been developed, with external diameters of approximately 2 and 1 mm. We have determined their important dosimetric characteristics (depth dose relation, angular dependence, energy dependence). We have shown that the background signal created by Cerenkov and fibre fluorescence does not significantly affect the performance in most clinical geometries. The dosimeter design enables readout at less than 0.5 s intervals. The clinical demands of real time in-vivo brachytherapy dosimetry can uniquely be satisfied by the BrachyFOD{sup TM}.

  12. Comparative study of some new EPR dosimeters

    International Nuclear Information System (INIS)

    Alzimami, K.S.; Maghraby, Ahmed M.; Bradley, D.A.

    2014-01-01

    Investigations have been made of four new radiation dosimetry EPR candidates from the same family of materials: sulfamic acid, sulfanillic acid, homotaurine, and taurine. Mass energy attenuation coefficients, mass stopping power values and the time dependence of the radiation induced radicals are compared. Also investigated are the microwave saturation behavior and the effect of applied modulation amplitude on both peak-to-peak line width (W PP ) and peak-to-peak signal height (H PP ). The dosimeters are characterized by simple spectra and stable radiation-induced radicals over reasonable durations, especially in taurine dosimeters. Sulfamic acid dosimeters possessed the highest sensitivity followed by taurine and homotaurine and sulfanillic. - Highlights: ► Several EPR dosimeters were suggested based on SO 3 − radical. ► Taurine, homotaurine, sulfanilic, and sulfamic acid all possess simple EPR spectra. ► Dosimeters were compared to each other in terms of the dosimetric point of view. ► Energy dependence curves of the selected dosimeters were compared to eachother

  13. Dosimeter

    International Nuclear Information System (INIS)

    Thomson, I.

    1986-01-01

    This invention relates to a dosimeter for measuring ionizing radiation, and particularly to a dosimeter using an insulated gate field effect transistor (IGFET) as a sensor, having substantially improved accuracy. An IGFET is a field effect transistor fabricated on a silicon substrate and having an oxide insulator between the gate electrode and the silicon substrate. The gate electrode can be either metal or polycrystalline silicon dioxide. This invention overcomes previously-noted problems with IGFET sensors - the variation of threshold voltage with temperature, their inherent zero offset which varies from wafer to wafer, and the zero drift in threshold voltage - by measuring the differential threshold between two IGFET sensors exposed to the same radiation, in which one is biased into its conducting region, and the other is biased either off or to a conducting level less than the first. The measured differential threshold voltage between the two transistors will be a measure of the gamma radiation dose

  14. TH-CD-201-08: Flexible Dosimeter Bands for Whole-Body Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Kim, T; Fahimian, B; Pratx, G [Department of Radiation Oncology, Stanford University, Palo Alto, CA (United States)

    2016-06-15

    Purpose: The two commonly used radiotherapy techniques are total body irradiation (TBI) and the total skin irradiation (TSI). In order to ensure the accuracy of the prescription beams, the dose received throughout the entire body must be checked using dosimetry. However, the available number of data points is limited as the dosimeters are manually placed on the patient. We developed a flexible and wearable dosimeter that can collect 1D continuous dose information around the peripheral of the patients’ body, including areas obscured from the beam path. Methods: The flexible dosimeter bands are fabricated by embedding storage phosphor powders in a thin layer of non-toxic silicone based elastomer (PDMS). An additional elastomer layer is formed on top of the phosphor layer to provide additional mechanical support for the dosimeter. Once the curing process is complete, the dosimeter is cut into multiple bands and rolled into spools prior to use. Results: The dose responses are tested using a preclinical cabinet X-ray system, where the readout is performed with a storage phosphor reader. Results show that the dose calibration factor is ∼1400 (A.U./Gy) from the beam center. Also, 1-D dose distribution experiment was performed in water phantoms, where preliminary results demonstrate that the dose in water is indeed attenuated compared to in air. Conclusion: Dose response and high-resolution 1-D dosimetry is demonstrated using the flexible dosimeters. By providing a detailed spatial description of the beam dose profile, we expect that the dosimeter bands may aid in enhancing the current existing modality in dosimetry. Since the dosimeter is flexible (can retract back to its original length), they can be comfortably worn around the patient. Potentially, multiple 1-D dose information can be stitched together and extrapolated to provide a coarse 3-D image of the dose distribution. This work was supported by funding from the Cutaneous Lymphoma Foundation under the CLARIONS

  15. TH-CD-201-08: Flexible Dosimeter Bands for Whole-Body Dosimetry

    International Nuclear Information System (INIS)

    Kim, T; Fahimian, B; Pratx, G

    2016-01-01

    Purpose: The two commonly used radiotherapy techniques are total body irradiation (TBI) and the total skin irradiation (TSI). In order to ensure the accuracy of the prescription beams, the dose received throughout the entire body must be checked using dosimetry. However, the available number of data points is limited as the dosimeters are manually placed on the patient. We developed a flexible and wearable dosimeter that can collect 1D continuous dose information around the peripheral of the patients’ body, including areas obscured from the beam path. Methods: The flexible dosimeter bands are fabricated by embedding storage phosphor powders in a thin layer of non-toxic silicone based elastomer (PDMS). An additional elastomer layer is formed on top of the phosphor layer to provide additional mechanical support for the dosimeter. Once the curing process is complete, the dosimeter is cut into multiple bands and rolled into spools prior to use. Results: The dose responses are tested using a preclinical cabinet X-ray system, where the readout is performed with a storage phosphor reader. Results show that the dose calibration factor is ∼1400 (A.U./Gy) from the beam center. Also, 1-D dose distribution experiment was performed in water phantoms, where preliminary results demonstrate that the dose in water is indeed attenuated compared to in air. Conclusion: Dose response and high-resolution 1-D dosimetry is demonstrated using the flexible dosimeters. By providing a detailed spatial description of the beam dose profile, we expect that the dosimeter bands may aid in enhancing the current existing modality in dosimetry. Since the dosimeter is flexible (can retract back to its original length), they can be comfortably worn around the patient. Potentially, multiple 1-D dose information can be stitched together and extrapolated to provide a coarse 3-D image of the dose distribution. This work was supported by funding from the Cutaneous Lymphoma Foundation under the CLARIONS

  16. Intercomparison measurements with albedo neutron dosimeters

    International Nuclear Information System (INIS)

    Alberts, W.G.; Kluge, H.

    1994-01-01

    Since the introduction of the albedo dosimeter as the official personal neutron dosimeter the dosimetry services concerned have participated in intercomparison measurements at the PTB. Their albedo dosimeters were irradiated in reference fields produced by unmoderated and D 2 O-moderated 252 Cf neutron sources in the standard irradiation facility of the PTB. Six fields with fluences different in energy and angle distribution could be realised in order to determine the response of the albedo dosimeter. The dose equivalent values evaluated by the services were compared with the reference values of the PTB for the directional dose equivalent H'(10). The results turned out to be essentially dependent on the evaluation method and the choice of the calibration factors. (orig.) [de

  17. Evaluation of LLNL's Nuclear Accident Dosimeters at the CALIBAN Reactor September 2010

    International Nuclear Information System (INIS)

    Hickman, D.P.; Wysong, A.R.; Heinrichs, D.P.; Wong, C.T.; Merritt, M.J.; Topper, J.D.; Gressmann, F.A.; Madden, D.J.

    2011-01-01

    The Lawrence Livermore National Laboratory uses neutron activation elements in a Panasonic TLD holder as a personnel nuclear accident dosimeter (PNAD). The LLNL PNAD has periodically been tested using a Cf-252 neutron source, however until 2009, it was more than 25 years since the PNAD has been tested against a source of neutrons that arise from a reactor generated neutron spectrum that simulates a criticality. In October 2009, LLNL participated in an intercomparison of nuclear accident dosimeters at the CEA Valduc Silene reactor (Hickman, et.al. 2010). In September 2010, LLNL participated in a second intercomparison of nuclear accident dosimeters at CEA Valduc. The reactor generated neutron irradiations for the 2010 exercise were performed at the Caliban reactor. The Caliban results are described in this report. The procedure for measuring the nuclear accident dosimeters in the event of an accident has a solid foundation based on many experimental results and comparisons. The entire process, from receiving the activated NADs to collecting and storing them after counting was executed successfully in a field based operation. Under normal conditions at LLNL, detectors are ready and available 24/7 to perform the necessary measurement of nuclear accident components. Likewise LLNL maintains processing laboratories that are separated from the areas where measurements occur, but contained within the same facility for easy movement from processing area to measurement area. In the event of a loss of LLNL permanent facilities, the Caliban and previous Silene exercises have demonstrated that LLNL can establish field operations that will very good nuclear accident dosimetry results. There are still several aspects of LLNL's nuclear accident dosimetry program that have not been tested or confirmed. For instance, LLNL's method for using of biological samples (blood and hair) has not been verified since the method was first developed in the 1980's. Because LLNL and the other DOE

  18. Application of solid dosimeter to radiation control

    International Nuclear Information System (INIS)

    Tsujimoto, Tadashi

    1988-01-01

    Individual exposure dose measuring devices are used to measure the dose of each person in facilities using radiations. Major devices of this type currently used in Japan include the film badge, thermoluminescence dosimeter, portable radiation dosimeter and fluorescent glass dosimeter. All of these devices except the portable radiation dosimeter are of a solid type. Various portable-type spatial dose rate measuring devices, generally called survey meters, are available to determine the spatial distribution of radiations. Major survey meters incorporates an ionization chamber, GM counter tube or scintillation counter, while BF 3 counting tubes are available for neutron measurement. Of these, the scintillation dosimeter is of a solid type. A new scintillation survey meter has recently been developed which incorporated a discrimination bias modulation circuit. Dosimeters incorporating an ionization chamber or a GM counter tube are generally used as portable alarms. Recently, a new solid-type alarm has been developed which incorporates a solicon radiation detector. Microcomputers are also used for self-diagnosis, data processing, automatic calibration, etc. (Nogami, K.)

  19. Is it really not possible to use electronic personal dosimeters in clinical exposure situations?

    International Nuclear Information System (INIS)

    Borowski, M.; Poppe, B.; Looe, H.K.; Boetticher, H. von

    2010-01-01

    Purpose: Due to significant measuring inaccuracies that can occur under certain conditions, the use of electronic personal dosimeters in statutory measurements in X-ray diagnostics is currently legally restricted. The present study investigates the clinically relevant situations in which measurement errors of more then 20 % can be anticipated. Materials and Methods: Four series of experiments were made, comparing the results of the electronic personal dosimeter EPD Mk2.3 to those of reference dosimeters (TLDs and diagnostic dosimeters). On the one hand, personal doses have been determined in the routine operation of controlled areas in various departments. On the other hand, measurements on phantoms have been conducted in extreme but realistic situations under radiation protection. Experiments were conducted in unweakened scattered radiation as well as in unattenuated and attenuated direct radiation. Results: The tested electronic personal dosimeter type meets the requirements regarding measurement accuracy for 'official' personal dosimeters in all of the examined clinically relevant scattered radiation fields. Only if exposed to radiation directly, an underestimation of the dose can occur and can be greater than 90 %. Conclusion: In the range of scattered radiation of diagnostic X-ray equipment, even in pulsed fields, the use of electronic personal dosimeters is reasonable. Considerable measurement errors can only arise in radiation fields that are not realistic under regular conditions and even in connection with most accidents. (orig.)

  20. Personnel neutron dosimeter evaluation and upgrade program

    International Nuclear Information System (INIS)

    Fix, J.J.; Brackenbush, L.W.; McDonald, J.C.; Roberson, P.L.; Holbrook, K.L.; Endres, G.W.R.; Faust, L.G.

    1983-01-01

    Evaluation of neutron dosimeters from twelve DOE laboratories involved about 2500 dosimeter irradiations at both PNL and the National Bureau of Standards (NBS) using neutrons of several energies and doses and several irradiations for good statistical analysis. The data and their analyses will be published later. The information evaluates accuracy, precision, lower dose detection, and energy response of dosimeters

  1. An environmental BeO-OSL dosimeter for emergency response

    International Nuclear Information System (INIS)

    Woda, Clemens; Kaiser, Jan Christian; Urso, Laura; Greiter, Matthias

    2012-01-01

    A conceptual design is presented to use measurements of localized absorbed dose in inner cities for production of high resolution maps of the radioactive contamination following a nuclear emergency or radiological attack. The doses are derived from luminescent detectors pre-fixed at places of high importance (e.g. public squares). For such an environmental dosimeter, BeO is used, which can be read out using optically stimulated luminescence (OSL). A suitable casing of black Perspex has been developed to give a sufficiently accurate estimate of the air kerma value at the detector position. The dosimeter is characterized according to light tightness, dose response and angular photon energy dependence. A short overview of the approach for map production is also given. - Highlights: ► An inexpensive, environmentally stable BeO based OSL dosimeter has been developed for emergency response. ► The detector enables fast readouts and shows highly favorable dosimetric properties. ► A conceptual design is described to produce maps of radioactive contamination from localized dose measurements in urban areas.

  2. Characterization of a new dosimeter for the development of a position-sensitive detector of radioactive sources in industrial NDT equipment

    Science.gov (United States)

    Kim, K. T.; Kim, J. H.; Han, M. J.; Heo, Y. J.; Park, S. K.

    2018-02-01

    Imaging technology based on gamma-ray sources has been extensively used in non-destructive testing (NDT) to detect any possible internal defects in products without changing their shapes or functions. However, such technology has been subject to increasingly stricter regulations, and an international radiation-safety management system has been recently established. Consequently, radiation source location in NDT systems has become an essential process, given that it can prevent radiation accidents. In this study, we focused on developing a monitoring system that can detect, in real time, the position of a radioactive source in the source guide tube of a projector. We fabricated a lead iodide (PbI2) dosimeter based on the particle-in-binder method, which has a high production yield and facilitates thickness and shape adjustment. Using a gamma-ray source, we then tested the reproducibility, linearity of the dosimeter response, and the dosimeter's percentage interval distance (PID). It was found that the fabricated PbI2 dosimeter yields highly accurate, reproducible, and linear dose measurements. The PID analysis—conducted to investigate the possibility of developing a monitoring system based on the proposed dosimeter—indicated that the valid detection distance was approximately 11.3 cm. The results of this study are expected to contribute to the development of an easily usable radiation monitoring system capable of significantly reducing the risk of radiation accidents.

  3. Energy-dependency correction factors for the digital dosimeters using in NMD environment dose assessment

    International Nuclear Information System (INIS)

    Lai, Y.C.; Huang, Y. F.; Chen, Y.W.

    2008-01-01

    Full text: Short-term environment dose-rate assessments using real-time digital dosimeters within a Nuclear Medicine Department (NMD) are gaining more world-wide uses recently. In the past, conventional ion chamber-type survey-meters are used dominantly in environmental dose rates evaluation. Although it has suffered less gamma energy-dependency, but it is less sensitive in comparison with other digital dosimeters and more bulky in design that can hardly make it into a pocket size application. With modern electronic advancement and its shrinking in physical size, real-time personal dosimeter nowadays has gaining more popular to use a miniature G-M counter or a solid-state diode sensor, or even a NaI(Tl) scintillation device for ambient radiation monitoring. Radiation sensor operated in pulse-mode can never been used in doses or dose rates determination since each digital pulse has carried no energy information of the impinging gamma ray being interactive with, especially in the G-M counter or the diode sensor case. The raw count rates measured from a pulse-mode device are heavily dependent on the packaging of the sensor to make it less energy-sensitive. The doses or dose rates are then calculated by using a built-in conversion factor, based on a Cs-137 beam source calibration data conducted by various manufacturing vendors, to convert its raw counts into a so-called dose or dose-rate unit. In this study, we have focused our interests in the low energy response of the digital dosimeters from several brands currently for our in-house uses. Mainly, Tc-99m and I-131 in point sources and water phantoms detection configurations have been deployed to simulate our NMD outpatients for environment radiation monitoring purpose. The energy-dependent correction factors of the digital dosimeters will be evaluated by using calibrated Tc-99m or I-131 standard sources directly that has much lower gamma energy than the Cs-137 beam source of 661 keV. In the near future, we would

  4. Effects of composition interactions on the response of a turnbull blue radiochromic gel dosimeter

    International Nuclear Information System (INIS)

    Shieh, Jiunn-I; Cheng, Kai-Yuan; Shyu, Huey-Lih; Yu, Yi-Chen; Hsieh, Ling-Ling

    2014-01-01

    In this study, the Taguchi statistical method was used to design experiments for investigating the effects of interactions among compositions on the performance of a Turnbull blue gel (TBG) radiochromic dosimeter. Four parameters were considered as the design factors: (A) concentration of ferric chloride, (B) concentration of potassium ferricyanide, (C) concentration of sulfuric acid, and (D) amount of gelling agent added. Two levels were selected for each factor. The change in optical absorbance at 695 nm under UVA exposures was monitored to determine the response of the dosimeters. The results showed that the contributions of factors A–D on the absorbance were 20.01%, 23.16%, 27.03%, and 0.49%, respectively. The contributions of significant interaction effects were AC (8.60%), BC (5.61%), and ABC (10.56%). This finding indicated that sulfuric acid (C) was the most influential factor, whereas gelling agent (D) was the least influential factor. Sulfuric acid had an important function in two two-way interactions and one three-way interaction in the response of TBG to UV exposure. - Highlights: • Analysis of the composition that influence TBG dosimeters via the design of experiments. • Cross interactions between factors in the TBG dosimeters through multi-factor ANOVA. • Two two-way interactions and one three-way interaction in the TBG dosimeters are significant

  5. Android Based Area Web Monitoring

    Science.gov (United States)

    Kanigoro, Bayu; Galih Salman, Afan; Moniaga, Jurike V.; Chandra, Eric; Rezky Chandra, Zein

    2014-03-01

    The research objective is to develop an application that can be used in the monitoring of an area by using a webcam. It aims to create a sense of security on the user's application because it can monitor an area using mobile phone anywhere. The results obtained in this study is to create an area with a webcam monitoring application that can be accessed anywhere as long as the monitoring results have internet access and can also be accessed through Android Based Mobile Phone.

  6. Android Based Area Web Monitoring

    Directory of Open Access Journals (Sweden)

    Kanigoro Bayu

    2014-03-01

    Full Text Available The research objective is to develop an application that can be used in the monitoring of an area by using a webcam. It aims to create a sense of security on the user's application because it can monitor an area using mobile phone anywhere. The results obtained in this study is to create an area with a webcam monitoring application that can be accessed anywhere as long as the monitoring results have internet access and can also be accessed through Android Based Mobile Phone.

  7. Mexican gems as thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Azorin N, J.

    1979-01-01

    The possibility of using naturally ocurring mexican gems as thermoluminescent dosimeters (TLD) was investigated. Twelve types of gems were irradiated with X and gamma rays in order to determinate their dosimetric properties. Three of these gems showed favorable thermoluminescent characteristics compared with commercial thermoluminescent dosimeters. The plots of their thermoluminescent response as a function of gamma dose are straight lines on full log paper in the dose range 10 -2 to 10 2 Gy. The energy dependence is very strong to low energies of the radiation. Their fading was found to be about 5%/yr. and they may be annealed as reused without loss in sensitivity. Therefore, these gems can be used as X and gamma radiation dosimeters. (author)

  8. Research on the formula of radiochromic film dosimeters

    International Nuclear Information System (INIS)

    Li Huazhi; Xiao Zhenhong; Lin Min; Cui Ying; Chen Kesheng; Chen Yundong; Ye Hongsheng; Lin Jingwen

    2006-10-01

    The formula of radiochromic film dosimeters was studied. Commercially available nylon was used as the matrix, while hexahydroxyethyl pararosaniline cyanide (HPR-CN) and pararosaniline cyanide (PR-CN) that are made in China and other countries were used as the dyes of the dosimeters. the performance of the thin film dosimeters made in CIAE was tested and compared with each other. The formula of the dosimeters was finally confirmed by testing its physical properties and dosimetric characteristics. (authors)

  9. Development and dosimetric evaluation of radiochromic PCDA vesicle gel dosimeters

    International Nuclear Information System (INIS)

    Sun, P.; Fu, Y.C.; Hu, J.; Hao, N.; Huang, W.; Jiang, B.

    2016-01-01

    The gel dosimeter has the unique capacity in recording radiation dose distribution in three dimensions (3D), which has the specific advantages in dosimetry measurements where steep dose gradients exist, such as in intensity-modulated radiation therapy (IMRT), brachytherapy and so on. Some 3D dosimeters, such as Fricke gel dosimeters, polymer gel dosimeters, the PRESAGE plastic dosimeters and micelle gel dosimeters have appeared recently. However, there are several disadvantages of these 3D dosimeters limit their application in radiotherapy dose verification. In this study, a novel radiochromic gel dosimeter for 3D dose verification of radiotherapy was developed by dispersing nanovesicles self-assembled by 10,12-pentacosadiynoic acid (PCDA) into the tissue equivalence gel matrix. The characteristics of radiochromic PCDA vesicle gel dosimeters were evaluated. The results indicate that these radiochromic gel dosimeters have good linear dose response to X-ray irradiation in the dose range of 2–100 Gy. In addition, the radiochromic gel dosimeters breakthrough the limitations of the existing gel dosimeters such as diffusion effect, post-radiation effect, and poor forming ability. The response of the gel dosimeter does not show any dose rate dependence, energy dependence and temperature effect, and there was no obvious difference in the gel response between single and cumulative dose of fractional irradiation. Hence, the radiochromic PCDA vesicle gel dosimeters developed in this study could be generally applied to 3D dose verification in radiotherapy. - Highlights: • A novel radiochromic gel dosimeter was developed by dispersing PCDA nanovesicles into the tissue equivalence gel matrix. • This nanovesicle overcomes the dose image blurring caused by the diffusion of monomer molecules. • This nanovesicle limits the polymer chain growth, so as to reduce the post-radiation effect. • The gel matrixes possess excellent tissue equivalence and elastic strength, which

  10. Development and validation of a protocol for field validation of passive dosimeters for ethylene oxide excursion limit monitoring

    Energy Technology Data Exchange (ETDEWEB)

    Puskar, M.A.; Szopinski, F.G.; Hecker, L.H. (Corporate Industrial Hygiene Laboratory, Abbott Laboratories, North Chicago, IL (USA))

    1991-04-01

    An exposure and analysis protocol is described for the field validation of passive dosimeters for ethylene oxide (EtO) excursion limit monitoring. The protocol calls for the use of a field exposure chamber with concurrent sampling using Tedlar air-sampling bags. The bags are analyzed immediately after sampling by gas chromatography with flame ionization detection (GC-FID). The chamber design allows all monitors to be exposed for the exact same time in the field. The sampling and analysis procedure not only determines the actual concentration of EtO present during the monitor's exposure but estimates if concentrations of EtO vary from point to point in the monitor array during the exposure. In chamber operation, the accuracy of the standard generator used to calibrate the GC-FID was independently verified in the field by the standard additions method. The sampling bias of the sampling train was determined to be -3.5% in the 2.4 ppm to 14.3 ppm concentration range. To estimate the stability of collected EtO samples in Tedlar bags, the rate of EtO loss in the bags was determined to be 0.011 ppm/hr at 2.57 ppm and 0.066 ppm/hr at 8.07 ppm. Sampling bias of the passive methods by additional EtO exposure of the monitors in the closed chamber after sampling and during purging was determined to be +1.5%. The Tedlar bag sampling method with subsequent GC-FID determination demonstrated a coefficient of variation of 1.8% at 2.43 ppm.

  11. Magnetic field dosimeter development

    International Nuclear Information System (INIS)

    Lemon, D.K.; Skorpik, J.R.; Eick, J.L.

    1980-09-01

    In recent years there has been increased concern over potential health hazards related to exposure of personnel to magnetic fields. If exposure standards are to be established, then a means for measuring magnetic field dose must be available. To meet this need, the Department of Energy has funded development of prototype dosimeters at the Battelle Pacific Northwest Laboratory. This manual reviews the principle of operation of the dosimeter and also contains step-by-step instructions for its operation

  12. Use of wrist albedo neutron dosimeters

    International Nuclear Information System (INIS)

    Hankins, D.E.

    1983-01-01

    We are developing a wrist dosimeter that can be used to measure the exposure at the wrist to x-rays, gamma rays, beta-particles, thermal neutrons and fast neutrons. It consists of a modified Hankins Type albedo neutron dosimeter and also contains three pieces of CR-39 plastic. ABS plastic in the form of an elongated hemisphere provides the beta and low energy x-ray shielding necessary to meet the requirement of depth dose measurements at 1 cm. The dosimeter has a beta window located in the side of the hemisphere oriented towards an object being held in the hands. A TLD 600 is positioned under the 1 cm thick ABS plastic and is used to measure the thermal neutron dose. At present we are using Velcro straps to hold the dosimeter on the inside of the wrist. 9 figures

  13. Evaluation of fading factor and self-dose for glass dosimeter and thermoluminescence dosimeter

    International Nuclear Information System (INIS)

    Yamasaki, T.; Yamanishi, H.; Miyake, H.; Komura, K.

    2000-01-01

    The glass dosimeter (GD) and thermoluminescence dosimeter (TLD) are both passive radiation detectors. They are often used for measuring environmental radiation. In order to measure low dose rate preciously, it is important to evaluate decreased dose due to fading and self-dose during the exposure period. We evaluate the fading factor and self-dose of thee passive detectors, GD and TLD. We select Ogoya tunnel for the experiment. The tunnel is suitable field for measuring faded dose and self-dose because it is low cosmic radiation. At the center of the tunnel, the intensity of cosmic ray is reduced to about 1/177 than the outside of the funnel. We prepared two sets of dosimeters. One set consists of five GDs, five TLDs and some pre-irradiated GDs and TLDs that are exposed to standard radiation of 4 mGy by Cs-137. These dosimeters are put in the 10 cm thick lead box in order to shield the terrestrial gamma ray. One set is located at the center of the tunnel and the other is the outside of the funnel. The dosimeters were exposed for ten months, from May 1998 to March 1999. After the exposure, the readers of dosimeters are carried into the funnel to read out the signals promptly as soon as taking out the dosimeters. As a result of the measurement, four kinds of data are taken for GD and TLD respectively. Assumed that the self-dose and cosmic ray are constant during exposure, the four independent unknown quantities, a self-dose a dose due to cosmic ray and a fading coefficient at the center of the tunnel and at the outside, are considered. Therefore four simultaneous equations should be obtained. From these examinations, the faded dose of GD is less than 1%, but that of TLD is about 16% during ten months. The coefficient for compensation of fading of GD and TLD is given as the half of the each value. At the outside of the tunnel, the measured dose rate of cosmic ray that can pass through the 10 cm lead is evaluated to be about 16 nGy/h by both detectors. The self

  14. Suitability of CR-39 dosimeters for personal dosimetry around CANDU reactors

    International Nuclear Information System (INIS)

    Cross, W.G.

    1992-08-01

    The capabilities and limitations of CR-39 damage track detectors have been evaluated for their use as personal neutron dosimeters around CANDU reactors. Since the energy response is a critical characteristic, the neutron energy spectra expected within CANDU containments were studied. In the boiler rooms, around the moderator cooling systems, and in most of the fueling machine vaults, the spectra vary considerably, but the majority of the dose is expected to be delivered by neutrons above 80 keV, the approximate threshold for electrochemically-etched CR-39 detectors. In the Pickering A fueling machine vault, and in areas in other stations to which neutrons from reactors have been multiply scattered, lower energy neutrons may be important. In nearly all areas where people work, it appears that working times will be limited by gamma rays rather than by neutrons. The characteristics of other neutron dosimeters - bubble and superheated drop detectors, albedo detectors, and Si real-time detectors - were also reviewed. For workers who typically receive neutron doses that are small compared with regulatory limits, CR-39 is the most suitable available dosimeter for demonstrating compliance. All single dosimeters have poor angular response over the range 0 to 180 degrees because of the shielding of the body. Albedo and Si detectors have particularly poor energy responses over the energy range of importance. Bubble and superheated drop detectors have the advantages of immediate readout and high sensitivity, but the disadvantages of inability to integrate doses over a long period, temperature dependence, very limited range and higher cost. (Author) (110 refs., 45 figs.)

  15. Evaluation of a LiF:Mg,Ti thermoluminescent ring dosimeter according to the IEC 62387:2012 standards

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Edyelle L.B.; Barros, Vinícius S.M. de; Asfora, Viviane K.; Khoury, Helen J., E-mail: vsmdbarros@gmail.com [Unversidade Federal de Pernambuco (UFPE), Recife, PE (Brazil). Departamento de Energia Nuclear

    2017-07-01

    This work shows results of type testing of a ring dosimeter system under new IEC 62387:2012. The personal dosimeter investigated in this work consists of a commercial one element plastic ring (RADOS) which contains an LiF:Mg,Ti. By applying requirements for statistical fluctuations and linearity, a minimum measurable dose in Hp(0.07) was established. Energy and angular dependence aided in determining energy correction factors and fading requirements were used to select the most appropriate preheat scheme. Type testing of passive radiation monitors performed in the Radiation Metrology Laboratory (LMRI-DEN/UFPE) of the Federal University of Pernambuco is a major step in Brazil for the independent evaluation of these dosimeters, currently not available in the country. (author)

  16. Feasibility of smartphone diaries and personal dosimeters to quantitatively study exposure to ultraviolet radiation in a small national sample.

    Science.gov (United States)

    Køster, Brian; Søndergaard, Jens; Nielsen, Jesper B; Allen, Martin; Bjerregaard, Mette; Olsen, Anja; Bentzen, Joan

    2015-09-01

    In 2007, a national skin cancer prevention campaign was launched to reduce the UV exposure of the Danish population. To improve campaign evaluation a questionnaire validation using UV-dosimeters was initiated. To show the feasibility of dosimeters for national representative studies and of smartphones as a data collection tool. Participants were sent a dosimeter which they wore for 7 days, received a short diary questionnaire by text message each day and subsequently a longer questionnaire. Correlation between responses from questionnaire, smartphone diaries and dosimeters were examined. This study shows a 99.5% return rate (n = 205) of the dosimeters by ordinary mail and high response-rates for a smartphone questionnaire dairy. Correlation coefficients for outdoor-time reported through smartphones and dosimeters as average by week 0.62 (0.39-0.77), P questionnaire and dosimeters were 0.42 (0.11-0.64), P = 0.008. The subjective perception of the weather was the only covariate significantly influencing questionnaire estimates of actual outdoor exposure. We showed that dosimeter studies are feasible in national settings and that smartphones are a useful tool for monitoring and collecting UV behavior data. We found diary data reported on a daily basis through smartphones more strongly associated with actual outdoor time than questionnaire data. Our results demonstrate tools and possible considerations for executing a UV behavior questionnaire validation. © 2015 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

  17. Neutron dosimeter utilizing CR-39

    International Nuclear Information System (INIS)

    Souza, H.V.C. de.

    1991-05-01

    A personnel neutron dosimeter has been developed with discretization in a wide range of energies of real interest, utilizing the CR-39 polymer, to detect recoil protons in the fast range, and alpha particles in the thermal and epithermal ranges, with possibility to be disposed in the IRD/CNEN's conventional film badge suport. They are presented, abstractly, the difficulties and importance of the neutron dosimetry, beyond the general objectives that motivated this work execution. The details of the materials utilized in the dosimeter confection, and the experimental methodology employed to obtain the performance curves are presented. The results about linearity response of the dosimeter with respect to equivalent dose, in a wide range of doses, and about the verified angular dependence are analysed. (author)

  18. Intercomparison of high energy neutron personnel dosimeters

    International Nuclear Information System (INIS)

    McDonald, J.C.; Akabani, G.; Loesch, R.M.

    1993-03-01

    An intercomparison of high-energy neutron personnel dosimeters was performed to evaluate the uniformity of the response characteristics of typical neutron dosimeters presently in use at US Department of Energy (DOE) accelerator facilities. It was necessary to perform an intercomparison because there are no national or international standards for high-energy neutron dosimetry. The testing that is presently under way for the Department of Energy Laboratory Accreditation Program (DOELAP) is limited to the use of neutron sources that range in energy from about 1 keV to 2 MeV. Therefore, the high-energy neutron dosimeters presently in use at DOE accelerator facilities are not being tested effectively. This intercomparison employed neutrons produced by the 9 Be(p,n) 9 B interaction at the University of Washington cyclotron, using 50-MeV protons. The resulting neutron energy spectrum extended to a maximum of approximately 50-MeV, with a mean energy of about 20-MeV. Intercomparison results for currently used dosimeters, including Nuclear Type A (NTA) film, thermoluminescent dosimeter (TLD)-albedo, and track-etch dosimeters (TEDs), indicated a wide variation in response to identical doses of high-energy neutrons. Results of this study will be discussed along with a description of plans for future work

  19. VALIDATION OF HANFORD PERSONNEL AND EXTREMITY DOSIMETERS IN PLUTONIUM ENVIRONMENTS

    Energy Technology Data Exchange (ETDEWEB)

    Scherpelz, Robert I.; Fix, John J.; Rathbone, Bruce A.

    2000-02-10

    A study was performed in the Plutonium Finishing Plant to assess the performance of Hanford personnel neutron dosimetry. The study was assessed whole body dosimetry and extremity dosimetry performance. For both parts of the study, the TEPC was used as the principle instrument for characterizing workplace neutron fields. In the whole body study, 12.7-cm-diameter TEPCs were used in ten different locations in the facility. TLD and TED personnel dosimeters were exposed on a water-filled phantom to enable a comparison of TEPC and dosimeter response. In the extremity study, 1.27-cm-diameter TEPCs were exposed inside the fingers of a gloveboxe glove. Extremity dosimeters were wrapped around the TEPCs. The glove was then exposed to six different cans of plutonium, simulating the exposure that a worker's fingers would receive in a glovebox. The comparison of TEPC-measured neutron dose equivalent to TLD-measured gamma dose equivalent provided neutron-to-gamma ratios that can be used to estimate the neutron dose equivalent received by a worker's finger based on the gamma readings of an extremity dosimeter. The study also utilized a Snoopy and detectors based on bubble technology for assessing neutron exposures, providing a comparison of the effectiveness of these instruments for workplace monitoring. The study concludes that the TLD component of the HCND performs adequately overall, with a positive bias of 30%, but exhibits excessive variability in individual results due to instabilities in the algorithm. The TED response was less variable but only 20% of the TEPC reference dose on average because of the low neutron energies involved. The neutron response of the HSD was more variable than the TLD component of the HCND and biased high by a factor of 8 overall due to its calibration to unmoderated 252Cf. The study recommends further work to correct instabilities in the HCND algorithm and to explore the potential shown by the bubble-based dosimeters.

  20. Wallac automatic alarm dosimeter type RAD21

    International Nuclear Information System (INIS)

    Burgess, P. H.; Iles, W.J.

    1980-02-01

    The Automatic Alarm Dosimeter type RAD 21 is a batterypowered personal dosemeter and exposure rate alarm monitor, designed to be worn on the body, covering an exposure range from 0.1 to 999.9 mR and has an audible alarm which can be pre-set over the range 1 mR h -1 to 250 mR h -1 . The instrument is designed to measure x- and γ radiation over the energy range 50 keV to 3 MeV. The facilities and controls, the radiation, electrical, environmental and mechanical characteristics, and the manual, have been evaluated. (U.K.)

  1. Pulse monitor for upper extremities dosimetry in nuclear medicine

    International Nuclear Information System (INIS)

    Cledison de Jesus, Cunha; Divanizia do Nascimento, Souza

    2006-01-01

    In the manipulation of radioactive materials in Nuclear Medicine service the body parts of workers that are more displayed to the ionizing radiation are hands, underarm and arm. Therefore is necessary to developing personal dosimeters to monitoring of easy reproduction and low cost with purpose to determine the doses level radiation received by the worker in these extremities. However thermoluminescent dosimeters do not provide an instantaneous exposure reading, they are suitable for personal dosimetry because of their following advantages: wide useful dose range, small physical size and no need for high voltage or cables, i.e. stand alone character. The aim of this work is to investigation of a new pulse monitor, that has been developing with thermoluminescent detectors of CaSO 4 :Dy (TLD) using a small plate of acrylic, perforated cardboard to deposit the TLD. This set was involved in plastic to protect from humidity and other harmful ambient factors; moreover, a bracelet was inserted, adaptable for any worker. During the preparation of the personal dosimeters to monitor exposure it was necessary to verify their effectiveness to use by workers in a nuclear medicine service. The monitors have been submitted to procedures of performance evaluations by several tests: badges homogeneity, reproducibility, linearity, low detection limit, auto-irradiation, dosimeters stability, verification of the residual T.L. signal, visible light effect on dosimeters, energetic and angular dependence and TLD answer by influence of a simulator during radiation. Was possible to verify the efficiency of such upper extremities dosimeters and were obtained satisfactory results within of the limits demanded in the described tests above to this type of personal dosimeters. (authors)

  2. EPR of gamma-irradiated polycrystalline alanine-in-glass dosimeter

    International Nuclear Information System (INIS)

    Al-Karmi, Anan M.; Morsy, M.A.

    2008-01-01

    This study attempts to overcome some of the reported discrepancies in alanine-EPR reproducibility that may be related to alanine dosimeter preparation and/or EPR spectrometer settings. The dosimeters were prepared by packing pure polycrystalline L-α-alanine directly as supplied by the manufacturer in glass tubes. This dosimeter production scheme avoids any possible contribution to the EPR signal from a binding material. The dosimeters were irradiated with gamma ray to low-dose ranges typical for medical therapy (0-20 Gy). Special attention has been paid to the study of minimum detectable dose, measurement repeatability and reproducibility, and post-irradiation stability. The dosimeter exhibited a linear dose response in the dose range from 0.1 to 20 Gy. These positive properties favor the polycrystalline alanine-in-glass tube as a radiation dosimeter

  3. Application of Glycine-TTC dosimeter in gamma radiation processing facility

    International Nuclear Information System (INIS)

    Shinde, S.H.; Mondal, S.; Kulkarni, M.S.

    2018-01-01

    Glycine-TTC dosimeter was found to have a useful dose range of 5 to 30 kGy using spectro-photometric read-out method. Potential use of this dosimeter was demonstrated by measuring dose-rate in gamma chamber GC 900. The aim of the present study was to verify the performance of this dosimeter in actual industrial processing conditions encountered in radiation processing facility such as Gamma Radiation Processing Plant for Spices (GRPPS), BRIT, Vashi. Accordingly, glycine-TTC dosimeters were irradiated along with routine dosimeter viz. ceric-cerous of GRPPS and reference standard dosimeter viz. alanine EPR

  4. Bronchial dosimeter for radon progeny

    Energy Technology Data Exchange (ETDEWEB)

    Cheung, T.K.; Yu, K.N.; Nikezic, D.; Haque, A.K.M.M. [City University of Hong Kong, Hong Kong (China); Vucic, D. [Faculty of Technology, University of Nis, Lescovac (Yugoslavia)

    2000-05-01

    Traditionally, assessments of the bronchial dose from radon progeny were carried out by measuring the unattached fraction (f{sub p}) of potential alpha energy concentration (PAEC), the total PAEC, activity median diameters (AMDs) and equilibrium factor, and then using dosimetric lung models. A breakthrough was proposed by Hopke et al. (1990) to use multiple metal wire screens to mimic the deposition properties of radon progeny in the nasal (N) and tracheobronchial (T-B) regions directly. In particular, they were successful in using four layers of 400-mesh wire screens with a face velocity of 12 cm s{sup -1} for the simulation of radon progeny deposition in the T-B region. Oberstedt and Vanmarcke (1995) carried out precise calibrations for the system, and named the system as the 'bronchial dosimeter'. Based on these, Yu and Guan (1998) proposed a portable bronchial dosimeter similar to a normal measurement system for radon progeny or PAEC and consisted of only a single sampler and employed only one 400-mesh wire screen and one filter. However, all these 'bronchial dosimeters' in fact only determined the fraction of potential alpha energy from radon progeny deposited in the T-B region, which required certain assumptions and calculations to further give the final bronchial dose. In the present work, a true 'bronchial dosimeter' was designed, which consisted of three 400-mesh wire screens and a filter. With a face velocity of 11 cm s{sup -1}, the deposition pattern on the wire screens was found to satisfactorily match the variation of the dose conversion factor (in the unit of mSv/WLM) with the size of radon progeny from 1 to 1000 nm. In this way, this bronchial dosimeter directly gave the bronchial dose from the alpha counts recorded on the wire-screens and the filter paper. With the development of this bronchial dosimeter, the present practice of 'dose estimation' from large-scale radon surveys can be replaced by large

  5. Compton effect thermally activated depolarization dosimeter

    Science.gov (United States)

    Moran, Paul R.

    1978-01-01

    A dosimetry technique for high-energy gamma radiation or X-radiation employs the Compton effect in conjunction with radiation-induced thermally activated depolarization phenomena. A dielectric material is disposed between two electrodes which are electrically short circuited to produce a dosimeter which is then exposed to the gamma or X radiation. The gamma or X-radiation impinging on the dosimeter interacts with the dielectric material directly or with the metal composing the electrode to produce Compton electrons which are emitted preferentially in the direction in which the radiation was traveling. A portion of these electrons becomes trapped in the dielectric material, consequently inducing a stable electrical polarization in the dielectric material. Subsequent heating of the exposed dosimeter to the point of onset of ionic conductivity with the electrodes still shorted through an ammeter causes the dielectric material to depolarize, and the depolarization signal so emitted can be measured and is proportional to the dose of radiation received by the dosimeter.

  6. Dosimetry service participation of CIEMAT in intercomparisons 2008-2010 for personal dosimeters EURADOS; Participacion del servicio de dosimetria del CIEMAT en las intercomparaciones EURADOS 2008-2010 para dosimetros personales

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez Jimenez, R.; Romero Gutierrez, A. M.; Lopez Moyano, J. L.

    2011-07-01

    Individual monitoring of workers exposed to ionizing radiation requires the use of personal dosimeters. EURADOS (European Radiation Dosimetry Group) recently organized three intercomparison exercises External Personal Dosimetry Services (EPDS): two for body dosimeters in 2008 and 2010 and one for extremity dosimeters in the year 2009.El paper shows and analyzes the results obtained by CIEMAT SDPE participation in all exercises.

  7. Area monitoring intelligent system - SIMA

    International Nuclear Information System (INIS)

    Bhoem, P.; Hisas, F.; Gelardi, G.

    1990-01-01

    The area monitoring intelligent system (SIMA) is an equipment to be used in radioprotection. SIMA has the function of monitoring the radiation levels of determined areas of the installations where radioactive materials are handled. (Author) [es

  8. Personal solar UVR exposure studies using a new miniature electronic dosimeter/datalogger

    International Nuclear Information System (INIS)

    Gies, H.P.; Roy, C.R.; Toomey, S.; Borland, R.; Dixon, H.

    1996-01-01

    Full text: The Australian Radiation Laboratory (ARL) has been involved in a number of different collaborative studies around Australia using polysulphone (PS) film dosimeters in the measurement of personal exposure to solar ultraviolet radiation (UVR) (Herlihy E et al, Photochem Photobiol 60:288-294, 1994; Gies HP et al, Photochem Photobiol 62:1015-1021, 1995). These PS dosimeters have proven useful in measuring cumulative exposures during different outdoor occupational and recreational activities. Recent advances in both UV photodiodes and miniature dataloggers have allowed the development of UV dosimeters which can log the incident UVR exposure with time (Diffey BL and Saunders PJ, Photochem Photobiol 61:615-618, 1995). This provides information on the variation of UVR exposure with time of day and also on the effect of variations in local environment, activity and possibly behaviour on exposure. A pilot study to examine the performance of the UV dosimeter/ datalogger and to refine a suitable questionnaire for a much larger follow up study was undertaken in 1995. Two of the new dosimeter/dataloggers were available and a third, less portable unit was used to monitor ambient solar UVR. The site chosen was a recreation park within 5 km of ARL where calibrated ambient UVR monitoring occurs. The UVR exposures of people undertaking outdoor activities were measured. The subjects wore the UV detector attached to clothing on their back, so that it was out of sight and therefore less likely to cause them to consciously modify their behaviour during their activity. On completion the subjects were asked a number of standard questions about their activities while wearing the UV dosimeters. Of particular interest was whether the data collected could be used to determine whether subjects have modified their behaviour when in the sun as a result of educational campaigns on sun exposure run by the Cancer Councils. Comparison of the subjects UV data record with the simultaneous

  9. Calibration and testing of the DMG gamma dosimeter

    International Nuclear Information System (INIS)

    Dolgirev, E.I.

    1987-01-01

    25-1000 nGy/h (2.5-1000 μrad/h) absorbed dose gamma dosimeter for measuring the efficient equivalent irradiation dose for population is developed. It has two subranges 1000 nGy/h and 250 nGy/h. Results of dosimeter calibration and testing are presented. The dosimeter error for both subranges is less than 10%

  10. CRRES dosimeter simulations

    International Nuclear Information System (INIS)

    Auchampaugh, G.; Cayton, T.

    1993-04-01

    Conflicting data have been obtained from electron instruments aboard CRRES (Combined Release and Radiation Effects Satellite). To gain insight and to help in the interpretation of the data, we have calculated electron- and proton-flux and dose response functions for the four domes of the CRRES dosimeters using the Los Alamos Monte Carlo radiation transport codes. The response functions were calculated for electron and proton energies representative of those present in the space radiation environment. We also calculated the response of the dosimeters to a model radiation environment for orbit 607, which occurred on April 1, 1991 and compared the results to the measured values. The electron and proton components of the radiation environment were calculated using the solar maximum versions of the AE8 and AP8 models, namely, AE8MAX and AP8MAX. To facilitate the second task, we wrote two FORTRAN programs (CRRESunderscoreSIMP for AP8MAX and CRRESunderscoreSIME for AE8MAX) to read in a standard CRRES data file and to produce a comparison file of the calculated and measured values for all four dosimeter domes.The FORTRAN code will be available to the Phillips Laboratory for their use in making comparisons to other orbital data

  11. Reactor Gamma Heat Measurements with Calorimeters and Thermoluminescence Dosimeters

    DEFF Research Database (Denmark)

    Haack, Karsten; Majborn, Benny

    1973-01-01

    Intercomparison measurements of reactor γ-ray heating were carried out with calorimeters and thermoluminescence dosimeters. Within the measurement uncertainties the two methods yield coincident results. In the actual measurement range thermoluminescence dosimeters are less accurate than calorimet......Intercomparison measurements of reactor γ-ray heating were carried out with calorimeters and thermoluminescence dosimeters. Within the measurement uncertainties the two methods yield coincident results. In the actual measurement range thermoluminescence dosimeters are less accurate than...... calorimeters, but possess advantages such as a small probe size and the possibility of making simultaneous measurements at many different positions. Hence, thermoluminescence dosimeters may constitute a valuable supplement to calorimeters for reactor γ-ray heating measurements....

  12. Role of gel dosimeters in boron neutron capture therapy

    International Nuclear Information System (INIS)

    Khajeali, Azim; Farajollahi, Ali Reza; Khodadadi, Roghayeh; Kasesaz, Yaser; Khalili, Assef

    2015-01-01

    Gel dosimeters have acquired a unique status in radiotherapy, especially with the advent of the new techniques in which there is a need for three-dimensional dose measurement with high spatial resolution. One of the techniques in which the use of gel dosimeters has drawn the attention of the researchers is the boron neutron capture therapy. Exploring the history of gel dosimeters, this paper sets out to study their role in the boron neutron capture therapy dosimetric process. - Highlights: • Gel dosimeters have been investigated. • Conventional dosimetric proses of BNCT has been investigated. • Role of gel dosimeters in BNCT has been investigated

  13. Polymer gel dosimeter based on itaconic acid

    International Nuclear Information System (INIS)

    Mattea, Facundo; Chacón, David

    2015-01-01

    A new polymeric dosimeter based on itaconic acid and N, N’-methylenebisacrylamide was studied. The preparation method, compositions of monomer and crosslinking agent and the presence of oxygen in the dosimetric system were analyzed. The resulting materials were irradiated with an X-ray tube at 158 cGy/min, 226 cGy min and 298 cGy/min with doses up to 1000 Gy. The dosimeters presented a linear response in the dose range 75–1000 Gy, sensitivities of 0.037 1/Gy at 298 cGy/min and an increase in the sensitivity with lower dose rates. One of the most relevant outcomes in this study was obtaining different monomer to crosslinker inclusion in the formed gel for the dosimeters where oxygen was purged during the preparation method. This effect has not been reported in other typical dosimeters and could be attributed to the large differences in the reactivity among these species. - Highlights: • A novel polymer gel dosimeters based on itaconic acid is presented and characterized. • The typical linear trend of the dose behavior in a specific dose range was found. • Different gel structures were formed when oxygen and an antioxidant were present. • Absorbed dose is univocally correlated with optic absorbance and Raman spectroscopy. • Itaconic acid appears as a reliable radiation dosimeter that may be further improved.

  14. Raman Spectroscopy of Irradiated Normoxic Polymethacrylic Acid Gel Dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Bong, Ji Hye; Kwon, Soo Il; Cho, Yu Ra; Park, Chae Hee; Park, Hyung Wook [Kyonggi University, Suwon (Korea, Republic of); Choi, Kyu Seok; Yu, Soo Chang [Kunsan National University, Gunsan (Korea, Republic of)

    2011-02-15

    A quantitative analysis of the decreasing rate of the monomer and increasing rate of the polymerization was made by monitoring radiation level increments using Raman spectroscopy within the therapeutic radiation range for a normoxic polymethacrylic acid gel dosimeter. The gel dosimeter was synthesized by stirring materials such as gelatin, distilled water, methacrylic acid, hydroquinone and tetrakis phosphonium chloride at 50 .deg. C, and the synthesized gel was contained in a 10- mm diameter and 32-mm high vial to conduct measurement. 24 hours after gel synthesis, it was irradiated from 0 Gy to 20 Gy by 2 Gy using a Co-60 radiotherapy unit. With use of the Cryo FE-SEM, structural changes in the 0 Gy and 10 Gy gel dosimeters were investigated. The Raman spectra were acquired using 532-nm laser as the excitation source. In accordance with fitting the changes in C-COOH stretching (801 cm{sup -1}), C=C stretching (1639 cm{sup -1}) and vinyl CH{sub 2} stretching (3114 cm{sup -1}) vibrational modes for monomer and CH{sub 2} bending vibrational mode (1451 cm{sup -1}) for polymer, sensitive parameter S for each mode was calculated. The values of S for monomer bands and polymer band were ranged in 6.0 ± 2.6 Gy and 7.2 ± 2.3 Gy, respectively, which shows a relatively good conformity of the decreasing rate of monomer and the increasing rate of polymerization within the range of error.

  15. Performance testing of personnel extremity dosimeters by Korean LiF: Mg, Cu, Na, Si TLD(KLT-300)

    International Nuclear Information System (INIS)

    Kim, J.L.; Lee, J.I.; Chang, S.Y.; Choi, H.S.; Lee, D.H.; Han, S.J.

    2005-01-01

    Full text: As the needs and opportunities for utilization of atomic energy and radiation are increasing, the related industries, medicines, environments are developing and the relevant organizations and companies are also becoming diverse. In result, the types and kinds of the radiation related to occupational environments are becoming diversified. For the whole body dosimeters, the methodology and criteria for the performance evaluation and safety regulations and laws have been prepared in some detail, but for the extremity dosimeters, those are not prepared yet in Korea. The extremity dosimeters are required when the extremity part of our body, such as hand, elbow, and arm below the elbow, the foot, knee, and leg below the knee are exposed to the radiation in specific work environments. The dosimeter irradiation conditions are clearly discriminated between the whole body exposure condition and the extremity exposure condition. By the investigation and analysis of the management status and dose evaluation methods of the extremity dosimeters for the local absorbed dose, the personnel monitoring system of the extremity dosimeter services in Korea can be diagnosed, and the performance testing criteria and procedures can be established. Therefore, this study presents the performance testing results of extremity dosimeters on the finger and arm/leg phantoms by the procedures recommended in the ANSI (American National Standard) N13.32 using KLT-300 TL materials (LiF:Mg,Cu,Nas,Si) which were developed in Korea Atomic Energy Research Institute (KAERI). The results show that the performance index for the two types of phantoms are sufficiently satisfied with the prescribed tolerance level in the all of the test categories listed in the ANSI N13.32. These results and procedures used in this study can be applicable for regulatory body to establish the standard criteria for acceptable performance and testing conditions for personnel extremity dosimeters services in the

  16. Operational aspects of the direct ion storage dosimeter system: 18 months of experience at CERN

    International Nuclear Information System (INIS)

    Carbonez, P.; Kotamaki, E.; Otto, Th.

    2006-01-01

    CERN, the European Organization for Nuclear Research, operates a dosimetry service for external exposure. The service monitors approximately 5000 Staff, scientific visitors and contractors personnel working on the organizations sites with personal dosimeters for personal dose equivalent (Hp(10), Hp(0.07)) from gamma, beta and neutron radiation. The dosimetry service is approved by the Swiss Federal Health Office, the competent authority for radiation protection. In 2004, the R.A.D.O.S. D.I.S. -1 dosemeter has been introduced to CERN as a gamma-beta dosemeter. Technical characteristics of this dosimeter, based on the direct ion storage technology, are high sensitivity, excellent linearity of the personal dose response with respect to radiation energy or dose, and long-term physical storage of personal dose-related information without the risk of fading. One important technical feature of the dosemeter is its 'instant reading' capability: the user himself can evaluate the received personal dose nondestructively on specific reader stations. This information is digitized, centralized by the CERN-wide network and stored in a database. The consequence of the 'instant reading' capability is a break with in the traditional organisation of a dosimetry service. The personal dosimeters are no longer exchanged periodically for evaluation, but a monthly value of personal dose is calculated from the readings initiated by the user. After a wearing period of one year, users are invited to exchange the dosimeter against a new, recently calibrated unit. The introduction of the D.I.S.-1 dosimeter has profoundly changed the type of work in CERN dosimetry service. Technical and laboratory work (development of film dosimeters, densitometric evaluation) have made place for computer-based procedures and database management. (authors)

  17. An Emergency Dosimeter for Neutrons

    Energy Technology Data Exchange (ETDEWEB)

    Braun, J; Nilsson, R

    1960-05-15

    A neutron dosimeter suitable for single emergency exposures is described. The dosimeter is furnished with detectors for thermal, epi-thermal and fast neutrons. This means that three of the constants by which the spectrum of the incident neutron flux is approximated, can be determined. The dose calculated from these approximated spectra is compared to the dose from spectra obtained in different standard spectra of types which may be expected in a radiation accident.

  18. Review of Fricke gel dosimeters

    International Nuclear Information System (INIS)

    Schreiner, L J

    2004-01-01

    The innovation of adding a gel matrix to the traditional Fricke dosimeter to stabilize geometric information established the field of gel dosimetry for radiation therapy. A discussion of Fricke gels provides an overview of the issues that determine the dose response of all gel dosimeters in general. In this paper we review some of the features of Fricke systems to illustrate these issues and, in addition, to motivate renewed clinical interest in Fricke gels

  19. Fiber-optic dosimeters for radiation therapy

    Science.gov (United States)

    Li, Enbang; Archer, James

    2017-10-01

    According to the figures provided by the World Health Organization, cancer is a leading cause of death worldwide, accounting for 8.8 million deaths in 2015. Radiation therapy, which uses x-rays to destroy or injure cancer cells, has become one of the most important modalities to treat the primary cancer or advanced cancer. The newly developed microbeam radiation therapy (MRT), which uses highly collimated, quasi-parallel arrays of x-ray microbeams (typically 50 μm wide and separated by 400 μm) produced by synchrotron sources, represents a new paradigm in radiotherapy and has shown great promise in pre-clinical studies on different animal models. Measurements of the absorbed dose distribution of microbeams are vitally important for clinical acceptance of MRT and for developing quality assurance systems for MRT, hence are a challenging and important task for radiation dosimetry. On the other hand, during the traditional LINAC based radiotherapy and breast cancer brachytherapy, skin dose measurements and treatment planning also require a high spatial resolution, tissue equivalent, on-line dosimeter that is both economical and highly reliable. Such a dosimeter currently does not exist and remains a challenge in the development of radiation dosimetry. High resolution, water equivalent, optical and passive x-ray dosimeters have been developed and constructed by using plastic scintillators and optical fibers. The dosimeters have peak edge-on spatial resolutions ranging from 50 to 500 microns in one dimension, with a 10 micron resolution dosimeter under development. The developed fiber-optic dosimeters have been test with both LINAC and synchrotron x-ray beams. This work demonstrates that water-equivalent and high spatial resolution radiation detection can be achieved with scintillators and optical fiber systems. Among other advantages, the developed fiber-optic probes are also passive, energy independent, and radiation hard.

  20. Silicon Diode Dosimeter for Fast Neutrons

    International Nuclear Information System (INIS)

    Svansson, L.; Widell, C.O.; Swedberg, P.; Wik, M.

    1968-11-01

    The change of the current-voltage characteristics of a small silicon diode is used as a measure of fast neutron dose in the Fast Neutron Dosimeter 5422. This change is permanent and therefore it is possible to integrate doses over a long period of time. Doses from some rad up to 1000 rad can be measured and the information stored is not destroyed during readout. Considerable research work in this field has previously been carried out by the Swedish Institute for National Defence in collaboration with the Institute of Semiconductor Research Stockholm. The present investigation has been made in order to establish the possibilities of the dosimeter for practical applications and to study the variations of important parameters as a function of the production process. In particular the following parameters have been studied: - dose sensitivity, - energy dependence; - fading effect; - temperature influence; - maximum measurable dose. In general one might conclude that the dosimeter 5422 well fulfills requirements usually specified for a dosimeter for field service. Temperature influence and fading effect are of little practical importance within the recommended range of measurement

  1. Silicon Diode Dosimeter for Fast Neutrons

    Energy Technology Data Exchange (ETDEWEB)

    Svansson, L; Widell, C O; Swedberg, P [The Inst. of Semiconductor Researc h, Stockholm (Sweden); Wik, M [The Swedish Institute for National Defence, Sun dbyberg (Sweden)

    1968-11-15

    The change of the current-voltage characteristics of a small silicon diode is used as a measure of fast neutron dose in the Fast Neutron Dosimeter 5422. This change is permanent and therefore it is possible to integrate doses over a long period of time. Doses from some rad up to 1000 rad can be measured and the information stored is not destroyed during readout. Considerable research work in this field has previously been carried out by the Swedish Institute for National Defence in collaboration with the Institute of Semiconductor Research Stockholm. The present investigation has been made in order to establish the possibilities of the dosimeter for practical applications and to study the variations of important parameters as a function of the production process. In particular the following parameters have been studied: - dose sensitivity, - energy dependence; - fading effect; - temperature influence; - maximum measurable dose. In general one might conclude that the dosimeter 5422 well fulfills requirements usually specified for a dosimeter for field service. Temperature influence and fading effect are of little practical importance within the recommended range of measurement.

  2. Do you know DIS? a novel passive individual dosimeterd. Direct Ion Storage dosimeter DIS-1 officially approved in Switzerland

    International Nuclear Information System (INIS)

    Fiechtner, A.; Wernli, C.

    2001-01-01

    For individual monitoring film and TLD are the most often used types of dosimeters. On a smaller scale phosphate glasses and detectors based on optically stimulated luminescence (OSL) are also in use. As a new addition to the list of available personnel dosimeters the direct ion storage (DIS) system became legally approved for the first time in Switzerland. At the Paul Scherrer Institute (PSI) the RADOS dosimetry system DIS-1 is in official use since beginning of 2001. (orig.) [de

  3. Investigation of self-indicating radiation personal dosimeter

    International Nuclear Information System (INIS)

    Xia Wen; Ye Honsheng; Lin Min; Xu Lijun; Chen Kesheng; Chen Yizhen

    2014-01-01

    A self-indicating radiation personal dosimeter was investigated using radiation sensitive material diacetylene monomer PCDA, which was a component of the polymerization system. The substrate material, solvent, sensitive material, solution temperature, thickness of film and the preparation method were studied. The dosimeter colour changes from white to blue when exposed 0.1-2.5 Gy, and the linearly dependent coefficient of the exposure response is 0.9998, the stability of absorbency in two weeks after exposure is testified well. It can be used as self-indicating radiation alert personal dosimeter. (authors)

  4. Clinical dosimeter based on diamond detector

    International Nuclear Information System (INIS)

    Chervjakov, A.M.; Ljalina, L.I.; Ljutina, G.J.; Khrunov, V.S.; Martynov, S.S.; Popov, S.A.

    2002-01-01

    Full text: Diamond detectors have found application in the relative dosimetry and their parameters have been described elsewhere. Today, the exclusive producer of the diamond detector is the Institute of Physical and Technical Problems, Russia, and exclusive dealer is the PTW-Freiburg. The main features of the diamond detector are good long time stability, suitable range of the energy dependence for photon and electron beams in clinical use, independence of the measured date from temperature and pressure. The high sensitivity per volume unit of the diamond detector (1500 times higher than ionization chamber) allowed using detectors with very small volume (1-5 mm 3 ) and rather simple electronics for ionization current registration. The new dosimeter consists of the diamond detector itself, 40 m registration cable, pre-amplifier, micro-processor block for data handling and absorbed dose calculation using the calibration factor of diamond detector in terms of absorbed dose to water. Dosimeter has the possibility to work with PC using standard RS-232 interface. The main features of the dosimeter are as follows: the range of dose rate measurements for photon, electron and proton beams is within 0.01-1.0 Gy/s; the energy ranges for photons are 0.08-25 MeV, and 4-25 MeV for electrons, with energy dependence no more than ±2%; the main uncertainty of the dose measurements is within ±2%; the pre-irradiation dose for diamond detector is no more than 10 Gy; the sensitive volume of the used diamond detectors is within 1-5 mm 3 ; the weight of the dosimeter no more than 2 kg. The new dosimeter was evaluated at the Central Research Institute of Roentgenology and Radiology, St. Petersburg, Russia to verify its performance. The dosimeter was used as a reference instrument for dose measurements at Cobalt-60 unit, SL75-5 and SL-20 linear accelerators and the test results have shown that the device have met the specifications. It is planned to produce dosimeter as serial device by

  5. Commissioning optically stimulated luminescence in vivo dosimeters for fast neutron therapy

    Energy Technology Data Exchange (ETDEWEB)

    Young, Lori A., E-mail: layoung@uw.edu; Sandison, George [Department of Radiation Oncology, University of Washington, Seattle, Washington 98115 (United States); Yang, Fei [Sylvester comprehensive Cancer Center, University of Miami, Miami, Florida 33124 (United States); Woodworth, Davis [Department of Physics, University of Reno, Reno, Nevada 89557 (United States); McCormick, Zephyr [Department of Physics, University of California, Santa Barbara, California 93106 (United States)

    2016-01-15

    Purpose: Clinical in vivo dosimeters intended for use with photon and electron therapies have not been utilized for fast neutron therapy because they are highly susceptible to neutron damage. The objective of this work was to determine if a commercial optically stimulated luminescence (OSL) in vivo dosimetry system could be adapted for use in fast neutron therapy. Methods: A 50.5 MeV fast neutron beam generated by a clinical neutron therapy cyclotron was used to irradiate carbon doped aluminum oxide (Al{sub 2}O{sub 3}:C) optically simulated luminescence dosimeters (OSLDs) in a solid water phantom under standard calibration conditions, 150 cm SAD, 1.7 cm depth, and 10.3 × 10.0 cm field size. OSLD fading and electron trap depletion studies were performed with the OSLDs irradiated with 20 and 50 cGy and monitored over a 24-h period to determine the optimal time for reading the dosimeters during calibration. Four OSLDs per group were calibrated over a clinical dose range of 0–150 cGy. Results: OSLD measurement uncertainties were lowered to within ±2%–3% of the expected dose by minimizing the effect of transient fading that occurs with neutron irradiation and maintaining individual calibration factors for each dosimeter. Dose dependent luminescence fading extended beyond the manufacturer’s recommended 10 min period for irradiation with photon or electron beams. To minimize OSL variances caused by inconsistent fading among dosimeters, the observed optimal time for reading the OSLDs postirradiation was between 30 and 90 min. No field size, wedge factor, or gantry angle dependencies were observed in the OSLDs irradiated by the studied fast neutron beam. Conclusions: Measurements demonstrated that uncertainties less than ±3% were attainable in OSLDs irradiated with fast neutrons under clinical conditions. Accuracy and precision comparable to clinical OSL measurements observed with photons can be achieved by maintaining individual OSLD calibration factors and

  6. The development of web monitoring digital area monitor

    International Nuclear Information System (INIS)

    Jung, Hoon-Jin; Lee, Jun-Hee; Namkoong, Phil; Lee, Dong-Hoon; Lee, Su-Hong; Kim, We-Su

    2005-01-01

    As CCTV and radiation area monitor have been used separately to date, there have existed inconveniences in managing the view images and radiation dose rates from them. Thus we became to develop the web monitoring digital area monitor which incorporated the existing two factors, CCTV and radiation area monitor, into one. As incorporated with digitalisation, this device will usually become linked with PC, so that the view image and radiation dose rate can be concurrently identified in convenience. These measured figures automatically become data-based on PC by SW program, and become displayed in various format. Moreover, they can be monitored in remote and real time basis in the internet environment. Its local unit uses the cost-effective GM tube and CMOS image sensor, and has the small LCD which directly indicates the measured dose rate. The image sensor is designed to be operated with pan and tilt motion, thus can eliminate the dead view zone. It is thought that the developed device at this time could make the radiation safety management of each work field be done with low cost-high efficiency manner, making role of the CCTV inspection system

  7. The development of web monitoring digital area monitor

    Energy Technology Data Exchange (ETDEWEB)

    Jung, Hoon-Jin; Lee, Jun-Hee; Namkoong, Phil; Lee, Dong-Hoon; Lee, Su-Hong; Kim, We-Su [Iljin Radiation Engineering Co., Seoul (Korea, Republic of)

    2005-11-15

    As CCTV and radiation area monitor have been used separately to date, there have existed inconveniences in managing the view images and radiation dose rates from them. Thus we became to develop the web monitoring digital area monitor which incorporated the existing two factors, CCTV and radiation area monitor, into one. As incorporated with digitalisation, this device will usually become linked with PC, so that the view image and radiation dose rate can be concurrently identified in convenience. These measured figures automatically become data-based on PC by SW program, and become displayed in various format. Moreover, they can be monitored in remote and real time basis in the internet environment. Its local unit uses the cost-effective GM tube and CMOS image sensor, and has the small LCD which directly indicates the measured dose rate. The image sensor is designed to be operated with pan and tilt motion, thus can eliminate the dead view zone. It is thought that the developed device at this time could make the radiation safety management of each work field be done with low cost-high efficiency manner, making role of the CCTV inspection system.

  8. Personal monitoring

    International Nuclear Information System (INIS)

    1995-01-01

    Sources of ionizing radiation have innumerable applications in the workplace. The potential exposures of the individual workers involved may need to be routinely monitored and records kept of their cumulative radiation doses. There are also occasions when it is necessary to retrospectively determine a dose which may have been received by a worker. This Module explains the basic terminology associated with personal monitoring and describes the principal types of dosimeters and other related techniques and their application in the workplace. The Manual will be of most benefit if it forms part of more comprehensive training or is supplemented by the advice of a qualified expert in radiation protection. Most of the dosimeters and techniques described in this Module can only be provided by qualified experts

  9. Dose intercomparison study involving Fricke, ethanol chlorobenzene, PMMA and alanine dosimeters

    International Nuclear Information System (INIS)

    Lanuza, L.G.; Cabalfin, E.G.; Kojima, T.; Tachibana, H.

    1999-01-01

    A dose intercomparison study was carried out between the Philippine Nuclear Research Institute (PNRI) and Takasaki Radiation Chemistry Research Establishment, Japan Atomic Energy Research Institute (JAERI) to determine reliability of the dosimetry systems being used by PNRI employing ethanol chlorobenzene (ECB), Fricke and alanine dosimeters. The Fricke and ECB dosimeters were prepared at PNRI while the alanine-polystyrene dosimeter was provided by JAERI. Fricke or ECB dosimeters were irradiated together with alanine at PNRI gamma irradiation facilities. Analyses of the Fricke and ECB dosimeters were performed at PNRI while alanine dosimeters were analyzed at JAERI. A comparison study between alanine and polymethylmethacrylate (PMMA, Radix RN15) dosimeters was also undertaken at JAERI. The dosimeters were irradiated together under different irradiation conditions using the gamma irradiation facilities of JAERI and Radia Industry Co. Ltd. (Japan). Evaluations of PMMA and alanine dosimeters were both performed at JAERI. Result of the dose intercomparison of PNRI with the International Atomic Energy Agency through the International Dose Assurance Service (IDAS) is also presented. (author)

  10. Dose intercomparison study involving Fricke, ethanol chlorobenzene, PMMA and alanine dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Lanuza, L G; Cabalfin, E G [Philippine Nuclear Research Institute, Quezon City (Philippines); Kojima, T; Tachibana, H [Takasaki Radiation Chemistry Research Establishment, Japan Atomic Energy Research institute, Takasaki (Japan)

    1999-03-01

    A dose intercomparison study was carried out between the Philippine Nuclear Research Institute (PNRI) and Takasaki Radiation Chemistry Research Establishment, Japan Atomic Energy Research Institute (JAERI) to determine reliability of the dosimetry systems being used by PNRI employing ethanol chlorobenzene (ECB), Fricke and alanine dosimeters. The Fricke and ECB dosimeters were prepared at PNRI while the alanine-polystyrene dosimeter was provided by JAERI. Fricke or ECB dosimeters were irradiated together with alanine at PNRI gamma irradiation facilities. Analyses of the Fricke and ECB dosimeters were performed at PNRI while alanine dosimeters were analyzed at JAERI. A comparison study between alanine and polymethylmethacrylate (PMMA, Radix RN15) dosimeters was also undertaken at JAERI. The dosimeters were irradiated together under different irradiation conditions using the gamma irradiation facilities of JAERI and Radia Industry Co. Ltd. (Japan). Evaluations of PMMA and alanine dosimeters were both performed at JAERI. Result of the dose intercomparison of PNRI with the International Atomic Energy Agency through the International Dose Assurance Service (IDAS) is also presented. (author) 8 refs, 3 figs, 4 tabs

  11. CVD diamond detectors and dosimeters

    International Nuclear Information System (INIS)

    Manfredotti, C.; Fizzotti, F.; LoGiudice, A.; Paolini, C.; Oliviero, P.; Vittone, E.; Torino Univ., Torino

    2002-01-01

    Natural diamond, because of its well-known properties of tissue-equivalence, has recorded a wide spreading use in radiotherapy planning with electron linear accelerators. Artificial diamond dosimeters, as obtained by Chemical Vapour Deposition (CVD) could be capable to offer the same performances and they can be prepared in different volumes and shapes. The dosimeter sensitivity per unit volume may be easily proved to be better than standard ionization microchamber. We have prepared in our laboratory CVD diamond microchamber (diamond tips) in emispherical shape with an external diameter of 200 μm, which can be used both as X-ray beam profilometers and as microdosimeters for small field applications like stereotaxy and also for in vivo applications. These dosimeters, which are obtained on a wire substrate that could be either metallic or SiC or even graphite, display good performances also as ion or synchrotron X-rays detectors

  12. Calibration of dosimeters at 80-120 keV electron irradiation

    DEFF Research Database (Denmark)

    Miller, A.; Helt-Hansen, J.

    to calibrate thin-film dosimeters (Risø B3 and alanine films) by irradiation at the 80–120 keV electron accelerators. This calibration was compared to a 10MeV calibration, and we show that the radiation response of the dosimeter materials (the radiation chemical yield) is constant at these irradiation energies....... However, dose gradients within the dosimeters, when it is irradiated at low electron energies,mean that calibration function here will depend on both irradiation energy and the required effective point of measurement of the dosimeter. These are general effects that apply to any dosimeter that has a non...

  13. Evaluation of optical fibres as gamma radiation dosimeter

    International Nuclear Information System (INIS)

    Bohra, Dinesh; Chaudhary, H.S.; Panwar, Lalit; Vaijapurkar, S.G.; Bhatnagar, P.K.; Dasgupta, K.

    2005-01-01

    Semiconductor base gamma and neutron sensors are the fastest and popular dosimeters and are in competition with Thermoluminescence (TL) and Radio photoluminescence (RPL) dosimeters. All over the world armed forces require a dosimeter which records cumulative doses of ionizing radiations from mcGy to 10 Gy and is readable repeatedly without loss of dose information. TL dosimeters do not meet the criteria and RPL dosimeter meet the expectations and are in use by armed forces. Technologists have used laser as an excitation source to stimulate the glass and have achieved success in recording gamma doses of occupational/accidental span (mcGy to 10 Gy). However synthesizing RPL glass batches with exactly same characteristics predoses is a difficult task. Silicon base phosphorous doped step index multimode optical fibre can be made in a significant quantity and large number of dosimeters from it can be achieved with uniform predose. The radiation induced transmission loss gives a measure of gamma dose which is cumulative, readable repeatedly without loss of information. Assorted composition, core dia optical fibres have been synthesized and evaluated for dose linearity, dose rate independence, fading, length optimization. Here in is described some results of recent experiments and sensitivities achieved. (author)

  14. Measurements of SNAC2 area dosimeters placed in different configurations around the PROSPERO reactor and comparison with TRIPOLI-4 calculations

    Energy Technology Data Exchange (ETDEWEB)

    Rousseau, G.; Chambru, L.; Authier, N. [CEA, Centre de Valduc, 21120 Is-sur-Tille, (France)

    2015-07-01

    In the context of criticality accident alarm system tests, several experiments were carried out in 2013 on the PROSPERO reactor to study the response to neutron and gamma of different devices and dosimeters, particularly on the SNAC2 dosimeter. This article presents the results of this criticality dosimeter in different configurations, and compares the experimental measurements with the results of calculation performed with the TRIPOLI-4 Monte-Carlo Neutral Particles transport code. PROSPERO is a metallic critical assembly managed by the Criticality, Neutron Science and Measurement Department located at the French CEA Research Center of Valduc. The core, surrounded by a reflector of depleted uranium, is composed of 2 horizontal cylindrical blocks made of a highly enriched uranium alloy which can be placed in contact, and of 4 depleted uranium control rods which allow the reactor to be driven. This reactor, placed in a cell 10 m x 8 m x 6 m high, with 1.4-meter-thick concrete walls, is used as a fast neutron spectrum source and is operated at stable power level in delayed critical state, which can vary from 3 mW to 3 kW. PROSPERO is extensively used for electronic hardening or to study the effect of the neutrons on various materials. The SNAC2 criticality dosimeter is a zone dosimeter allowing the off line measurement of criticality accident neutron doses. This dosimeter consists of the pile up of seven activation foils embedded into a 23 mm diameter x 21 mm height cadmium container. The activation measurement of each foil, using a gamma spectroscopy technique, gives information about the neutron reaction rates. The SNAC2 software allows the spectrum unfolding from these values, taking into account the hypothesis of a particular spectrum shape, in three components: a Maxwell spectrum component for the thermal range, a 1/E component for the epithermal range, and a Watt spectrum component for the high energy range. Moreover, from the neutron spectrum, the SNAC

  15. Practical consequences for the use of a personal dosimeter for fast neutrons based on CR39 exposed up to one year

    International Nuclear Information System (INIS)

    Boschung, Markus; Fiechtner, Annette; Mayer, Sabine; Wernli, Christian

    2008-01-01

    Full text: At the Paul Scherrer Institut a personal neutron dosimetry system based on chemically etched CR-39 detectors and automatic track counting is in routine use since 1998. In its original design, the dosimeter is sensitive to thermal neutrons and to neutrons in the energy range from 200 keV up to several MeV. The standard exposition period is 3 months. Recently, a novel concept for individual monitoring was implemented at CERN. In this concept, each worker who possibly enters a radiation zone is equipped with a combined dosimeter for the measurement of personal photon and neutron doses. The dosimeter for photon dose measurement has an instant readout capability and dose measurements are done monthly. The dosimeter for neutron measurement is based on CR-39 detectors and is sensitive to fast neutrons only. The CR-39 detector is only evaluated and a neutron dose determined if the monthly personal photon dose exceeds 2 mSv or if the exposition period of the neutron dosimeter exceeds one year. This novel regime of use of the neutron dosimeter has had some important consequences for its practical implementation. A priori, the wearing period of a neutron dosimeter is not known and can range from 1 month up to 12 or even more months. A good knowledge of the long-term behaviour and characteristics of the detector material is needed. But also organisational and administrative issues have to be considered. The paper will outline the adopted procedure covering not only technical but also organisational aspects. The long-term behaviour of background track density and response to 241 Am-Be over one year are described as well as calibrations performed with 241 Am-Be and 252 Cf sources and in the High-Energy Reference Field Facility at CERN (CERF). The concept of individual monitoring at CERN could be transferred to other locations with high energy accelerators such as PSI and DESY. The experience gained with the neutron dosimeter based on CR-39 since introduction of the

  16. Limitations of commonly used thick-element personal dosimeters

    International Nuclear Information System (INIS)

    Gupta, V.P.

    1983-01-01

    In the ANSI Standard N13.11, accepted in June 1982, radiation dose depths of 1.0 cm and 0.007 cm in tissue for protection dosimetry have been adopted for deep and shallow dose equivalent estimations respectively. This standard is presently used for a mandatory personnel dosimetry performance testing program in the United States. Estimation of shallow-dose equivalent using a two-element dosimeter is described under the guidelines of this standard and the dosimetry practices followed by most dosimeter processors. A mathematical formulation, correlating a dosimeter response and shallow-dose equivalent factors at different energies, is presented. Also, the performance of a two-element thermoluminescent dosimeter is examined and the shallow-dose equivalent response results, both for the beta particles and photons, are discussed

  17. Improving the Success Rate of Delivering Annual Occupational Dosimetry Reports to Persons Issued Temporary External Dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Mallett, Michael Wesley [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2014-09-09

    Workers who are not routinely monitored for occupational radiation exposure at LANL may be issued temporary dosimeters in the field. Per 10CFR835 and DOE O 231.1A, the Laboratory's radiation protection program is responsible for reporting these results to the worker at the end of the year. To do so, the identity of the worker and their mailing address must be recorded by the delegated person at the time the dosimeter is issued. Historically, this data has not been consistently captured. A new online application was developed to record the issue of temporary dosimeters. The process flow of the application was structured such that: 1) the worker must be uniquely identified in the Lab's HR database, and 2) the mailing address of record is verified live time via a commercial web service, for the transaction to be completed. A COPQ savings (Type B1) of $96K/year is demonstrated for the new application.

  18. Monitoring of eye lens doses in radiation protection

    International Nuclear Information System (INIS)

    Bordy, J.M.

    2015-01-01

    Mainly due to the ICRP recommendation to decrease the exposure limit for eye lenses, the eye lens dosimetry has to be reconsidered. This paper gives an overview of the issues raised after this recommendation; that is to say, the choice and definition of the operational quantity to be monitored, the type testing and calibration of dosimeters aimed at measuring eyes lens 'doses', the design of existing eye lens dosimeters and their wearing conditions. Finally, a criterion to choose between a direct measurement of the personal dose equivalent at three millimeters depth, H p (3), with a dedicated dosimeter, and an indirect evaluation of H p (3) through whole-body monitoring is presented. (authors)

  19. Radiation detectors as surveillance monitors

    International Nuclear Information System (INIS)

    Fehlau, P.E.; Dowdy, E.J.

    1981-01-01

    The International Atomic Energy Agency (IAEA) proposes to use personnel dosimetry radiation detectors as surveillance monitors for safeguards purposes. It plans to place these YES/NO monitors at barrier penetration points declared closed under IAEA safeguards to detect the passage of plutonium-bearing nuclear material, usually spent fuel. For this application, commercially available dosimeters were surveyed as well as other radiation detectors that appeared suitable and likely to be marketed in the near future. No primary advantage was found in a particular detector type because in this application backgrounds vary during long counting intervals. Secondary considerations specify that the monitor be inexpensive and easy to tamper-proof, interrogate, and maintain. On this basis radiophotoluminescent, thermoluminescent, and electronic dosimeters were selected as possible routine monitors; the latter two may prove useful for data-base acquisition

  20. Intercomparison of personal dose equivalent measurements by active personal dosimeters. Final report of a joint IAEA-EURADOS project

    International Nuclear Information System (INIS)

    2007-11-01

    Active personal dosimeters (APD) are widely used in many countries, i.e. in the medical field and as operational dosimeters in nuclear power plants. Their use as legal dosimeters is already established in a few countries, and will increase in the near future. In the majority of countries, APDs have not undergone accreditation programmes or intercomparisons. In 2001, an EURADOS (European Radiation Dosimetry Group) Working Group on harmonization of individual monitoring was formed, funded by the European Commission, in the fifth framework programme, and by the participating institutes. The work addressed four issues; inter alia also an inventory of new developments in individual monitoring with an emphasis on the possibilities and performance of active (electronic) dosimeters for both photon/beta and neutron dosimetry. Within the work on this issue, a catalogue of the most extensively used active personal dosimeters (APDs) suitable for individual monitoring was made. On the basis of the knowledge gained in this activity, the organization of an international intercomparison, which would address APDs, was considered of great value to the dosimetric community. The IAEA in cooperation with EURADOS organized such an intercomparison in which most of the testing criteria as described in two internationally accepted standards (IEC61526 and IEC61283) were used. Additionally, simulated workplace fields were used for testing the APD reactions to pulsed X ray fields and mixed gamma/X ray fields. This is the first time that results of comparisons of such types are published, which is of great importance for APD end users in medical diagnostic and surgery X ray applications. Nine suppliers from six countries in Europe and the USA participated in the intercomparison with 13 different models. One of the models was a special design for extremity dose measurements. Irradiations and readout was done by two accredited calibration laboratories in Belgium and France and the French

  1. Solid-state personal dosimeter using dose conversion algorithm

    International Nuclear Information System (INIS)

    Lee, B.J.; Lee, Wanno; Cho, Gyuseong; Chang, S.Y.; Rho, S.R.

    2003-01-01

    Solid-state personal dosimeters using semiconductor detectors have been widely used because of their simplicity and real time operation. In this paper, a personal dosimeter based on a silicon PIN photodiode has been optimally designed by the Monte Carlo method and also developed. For performance test, the developed dosimeter was irradiated within the energy range between 50 keV and 1.25 MeV, the exposure dose rate between 3 mR/h and 25 R/h. The thickness of 0.2 mm Cu and 1.0 mm Al was selected as an optimal filter by simulation results. For minimizing the non-linear sensitivity on energy, dose conversion algorithm was presented, which was able to consider pulse number as well as pulse amplitude related to absorbed energies. The sensitivities of dosimeters developed by the proposed algorithm and the conventional method were compared and analyzed in detail. When dose conversion algorithm was used, the linearity of sensitivity was better about 38%. This dosimeter will be used for above 65 keV within the relative response of ±10% to 137 Cs

  2. An implantable radiation dosimeter for use in external beam radiation therapy

    International Nuclear Information System (INIS)

    Scarantino, Charles W.; Ruslander, David M.; Rini, Christopher J.; Mann, Gregory G.; Nagle, H. Troy; Black, Robert D.

    2004-01-01

    An implantable radiation dosimeter for use with external beam therapy has been developed and tested both in vitro and in canines. The device uses a MOSFET dosimeter and is polled telemetrically every day during the course of therapy. The device is designed for permanent implantation and also acts as a radiographic fiducial marker. Ten dogs (companion animals) that presented with spontaneous, malignant tumors were enrolled in the study and received an implant in the tumor CTV. Three dogs received an additional implant in collateral normal tissue. Radiation therapy plans were created for the animals and they were treated with roughly 300 cGy daily fractions until completion of the prescribed cumulative dose. The primary endpoints of the study were to record any adverse events due to sensor placement and to monitor any movement away from the point of placement. No adverse events were recorded. Unacceptable device migration was experienced in two subjects and a retention mechanism was developed to prevent movement in the future. Daily dose readings were successfully acquired in all subjects. A rigorous in vitro calibration methodology has been developed to ensure that the implanted devices maintain an accuracy of ±3.5% relative to an ionization chamber standard. The authors believe that an implantable radiation dosimeter is a practical and powerful tool that fosters individualized patient QA on a daily basis

  3. Evaluation of the implementation and use of active personal dosimeters for neutrons in Brazil; Evaluacion de la implementacion y del uso de los dosimetros individuales activos para neutrones en el Brasil

    Energy Technology Data Exchange (ETDEWEB)

    Castro B, C. P.; Wagner P, W.; De Souza P, K. C., E-mail: karla@ird.gov.br [Instituto de Radioprotecao e Dosimetria, Av. Salvador Allende s/n, Recreio Bandeirantes, 22780-160 Rio de Janeiro (Brazil)

    2014-08-15

    This work was conducted through of a field research based on a questionnaire sent to users of active personal dosimeters. A retrospective study of the last six years was also carried out of the services in the Neutron Metrology Laboratory (2008-2013) referent to the active personal dosimeters, taking into consideration the standards ISO-8529-3 and IEC-61526. The active personal dosimeters are defined as any instrument of individual monitoring with direct reading capacity, used by individuals exposed to ionizing radiation fields. Through research was verified that the active personal dosimeters work associated with other dosimeter types. Considering all dosimeters declared in the questionnaire, only two dosimeters (MGP brand Dmc 2000-GN model and the brand ATOMTEX model AT2503A) have conformity declaration with the international standard IEC-61526: 2005 reported by the manufacturers. (author)

  4. International intercomparison of environmental dosimeters under field and laboratory conditions

    International Nuclear Information System (INIS)

    Gesell, T.F.; de Planque Burke, G.; Becker, K.

    1975-04-01

    Based on the results of a pilot study at ORNL in 1973, a more comprehensive international intercomparison of integrating dosimeters for the assessment of external penetrating environmental radiation fields was carried out. Forty-one laboratories from eleven countries participated in this study. A total of 56 sets of six detectors each were mailed to and from Houston, Texas, where they were exposed for three months (July to September 1974) as follows: two in an unprotected space out-of-doors 1 m above ground; two in an air-conditioned shielded area with a known, low exposure rate; and two with the second group, but with an additional exposure to 30 mR. Evaluation of the dosimeters provides information on the calibration precision, the accuracy of field measurement, and transit exposure. Results are discussed. (U.S.)

  5. A unique radiation area monitoring system

    International Nuclear Information System (INIS)

    Murphy, P.C.; Allen, G.C.

    1978-01-01

    The Remote Area Monitoring Systems (RAMS) monitors four radiation areas with two independent systems in each area. Each system consists of power supplies, four ionization chambers, and four analog and digital circuits. The first system controls the warning beacons, horns, annunciation panel and interlocks. The second system presents a quantitative dose rate indication at the console and in the radiation area

  6. A PC based thin film dosimeter system

    DEFF Research Database (Denmark)

    Miller, A.; Hargittai, P.; Kovacs, A.

    2000-01-01

    A dosimeter system based on the Riso B3 dosimeter film, an office scanner for use with PC and the associated software is presented. The scanned image is analyzed either with standard software (Paint Shop Pro 5 or Excel) functions or with the computer code "Scanalizer" that allows presentation...

  7. Radiation protection instrumentation for personnel dosimetry, area and environmental monitoring

    International Nuclear Information System (INIS)

    Jones, A.R.

    1978-04-01

    Several kinds of personal dosimeter exist and their performance is reviewed in the light of requirements for an ideal instrument. The requirements of portable instruments are reviewed and the extent to which they are met in one example is described. Where permitted environmental doses are larger than the fluctuations in natural backgrounds, certain types of thermoluminescent dosimeters provide a cheaper and reliable alternative measuring system

  8. First assessment of individual monitoring of medical workers occupationally exposed to ionizing radiation in Burkina Faso

    International Nuclear Information System (INIS)

    Yakoro, A.; Nobila Ouédraogo, Salimata Traoré

    2010-01-01

    This paper reports the results of monitoring of medical workers occupationally exposed to ionizing radiation as a consequence of exposure to X-rays, from 2007 to 2010, in Burkina Faso. The radiation exposure monitoring was made with thermoluminescent dosimeters (TLD-100) type 0110 and the reader used was Harshaw 4500. The medical establishments. subscribers were provided with personal dosimeters (measuring Hp (10) and Hp (0.07)) and dosimeters for background and workplace exposure (H*10) measurement. The dosimeters have been worn for periods of 2 months each. The number of establishments subscribed and workers monitored has gradually increased from 4 radiology establishments with 13 workers monitored at September 2007 to 23 subscribers with 121 workers monitored at the end of April 2010. 13 establishments were still working without monitoring. From September 2007 to April 2010, no individual annual dose limit has really been reached. 88.16% of the 2 months dose values of personal dosimeters were below 0.1mSv, the detection limit and 96.61% of Hp (10) bimonthly values were below 3.33mSv. The workplace exposure monitoring values were often low (varying from 0.00mSv to 40.45mSv). 87.08% of the values of H*(10) were below 3.33mSv, the upper limit of Hp (10) for a period of 2 months. Low values of individual dose have also been recorded despite of high values of workplace monitoring. This allowed to state that the workers monitored were not exposed to a major risk. Nevertheless, 13 TLD have been lost and 3 damaged by subscribers (out of 1504 TLD provided). 26 times (out of 240), background measurement and workplace exposure monitoring dosimeters have been placed at the improper location. Therefore, sensitization of the establishments using ionizing radiation should be reinforced and the national regulations should impose radiation monitoring (author)

  9. Organic liquids as ''activ media'' in a holographic ionizing radiation dosimeter

    International Nuclear Information System (INIS)

    Nicolau-Rebigan, S.

    1979-01-01

    Some types of organic liquids for using as activ media in a holographic ionizing radiation dosimeter are presented. One outlined the advantages of the holographic dosimeter comparatively with those of common used dosimeters. One presented the advantages of utilization of the organic liquids comparatively with another chemical systems used in a holographic ionizing radiation dosimeter. (author)

  10. Radiation measured for ISS-Expedition 12 with different dosimeters

    International Nuclear Information System (INIS)

    Zhou, D.; Semones, E.; Gaza, R.; Johnson, S.; Zapp, N.; Weyland, M.

    2007-01-01

    Radiation in low Earth orbit (LEO) is mainly from Galactic Cosmic Rays (GCR), solar energetic particles and particles in South Atlantic Anomaly (SAA). These particles' radiation impact to astronauts depends strongly on the particles' linear energy transfer (LET) and is dominated by high LET radiation. It is important to investigate the LET spectrum for the radiation field and the influence of radiation on astronauts. At present, the best active dosimeters used for all LET are the tissue equivalent proportional counter (TEPC) and silicon detectors; the best passive dosimeters are thermoluminescence dosimeters (TLDs) or optically stimulated luminescence dosimeters (OSLDs) for low LET and CR-39 plastic nuclear track detectors (PNTDs) for high LET. TEPC, CR-39 PNTDs, TLDs and OSLDs were used to investigate the radiation for space mission Expedition 12 (ISS-11S) in LEO. LET spectra and radiation quantities (fluence, absorbed dose, dose equivalent and quality factor) were measured for the mission with these different dosimeters. This paper introduces the operation principles for these dosimeters, describes the method to combine the results measured by CR-39 PNTDs and TLDs/OSLDs, presents the experimental LET spectra and the radiation quantities

  11. Investigating On Colour Stability Conditions Of Postirradiation Radiochromic Film Dosimeter

    International Nuclear Information System (INIS)

    Nguyen Nguyet Dieu; Doan Binh; Pham Thu Hong; Cao Van Chung; Nguyen Thanh Duoc

    2011-01-01

    B3 dosimeter is a thin film with average thickness of 0.0194 mm, which is supplied by the Gex company, the United States. This dosimeter was influenced by many factors: light, temperature, humidity during and after irradiation process. In fact, B3 film dosimeters will be stable under certain conditions such as tightly sealed packs, controlled irradiation and stored temperature after irradiated. Therefore, investigation of the stability effect of postirradiated B3 film dosimeters on the heating temperature, heating time and storing time is carried out before the absorbed dose is read and followed standard reading procedures. When exposed to ionizing radiation, the dosimeters change from colorless to colour. The absorbed doses are read on a Genesys 20 spectrophotometer at a wavelength of 544 nm. Absorbed dose range is investigated from 0.55 to 80 kGy. Experimental results were indicated that colour stability of the postirradiated dosimeters at a temperature of 65 ± 3 o C for 30 minutes and keeping them in desiccator for 5 minutes before read out. Under these conditions, colour stability of B3 film dosimeter has maintained for 3 months. (author)

  12. SU-E-T-749: Thorough Calibration of MOSFET Dosimeters

    International Nuclear Information System (INIS)

    Plenkovich, D; Thomas, J

    2015-01-01

    Purpose: To improve the accuracy of the MOSFET calibration procedure by performing the measurement several times and calculating the average value of the calibration factor for various photon and electron energies. Methods: The output of three photon and six electron beams of Varian Trilogy linear accelerator SN 5878 was calibrated. Five reinforced standard sensitivity MOSFET dosimeters were placed in the calibration jig and connected to the Reader Module. As the backscatter material was used 7 cm of Virtual Water. The MOSFET dosimeters were covered with 1.5 cm thick bolus for the regular and SRS 6 MV beams, 3 cm bolus for 15 MV beam, 1.5 cm bolus for 6 MeV electron beam, and 2 cm bolus for the electron energies of 9, 12, 15, 18, and 22 MeV. The dosimeters were exposed to 100 MU, and the calibration factor was determined using the mobileMOSFET software. To improve the accuracy of calibration, this procedure was repeated ten times and the calibration factors were averaged. Results: As the number of calibrations was increasing the variability of calibration factors of different dosimeters was decreasing. After ten calibrations, the calibration factors for all five dosimeters were within 1% of one another for all energies, except 6 MV SRS photons and 6 MeV electrons, for which the variability was 2%. Conclusions: The described process results in calibration factors which are almost independent of modality or energy. Once calibrated, the dosimeters may be used for in-vivo dosimetry or for daily verification of the beam output. Measurement of the radiation dose under bolus and scatter to the eye are examples of frequent use of calibrated MOSFET dosimeters. The calibration factor determined for full build-up is used under these circumstances. To the best of our knowledge, such thorough procedure for calibrating MOSFET dosimeters has not been reported previously. Best Medical Canada provided MOSFET dosimeters for this project

  13. Application of the ALARA principle to individual occupational radiation monitoring

    International Nuclear Information System (INIS)

    Aleinikov, V.E.

    1998-01-01

    Routine individual occupational radiation monitoring is necessary to ensure that exposure are kept as low as reasonably achievable (ALARA) and that the authorized dose limits are not exceeded. The type and extend of the monitoring depend on the nature and magnitude of occupational radiation exposure. Optimal choice of the wearing time for individual dosimeters, T, has important consequences for achievement of the individual monitoring objects. The choice of T depends on individual dose distribution, the lower detection limit for the detector used, the lost of an information during period of monitoring, the cost of the measurements and the benefits associated with the results. The paper describes the application of quantitative optimization techniques to optimize the wearing time for individual dosimeters. It was shown how optimal wearing time for individual dosimeters depends on dose distribution, the lower detection limit, dose limit, fading and cost of the measurements. (author)

  14. The measurement of thoron concentrations in Australia using the Japanese passive R-T dosimeter

    International Nuclear Information System (INIS)

    Toussaint, L.F.; Tokonami, S.; Doi, M.; Solomon, S.B.; Peggie, J.R.

    1998-01-01

    Previous studies have shown higher gamma radiation levels and higher soil thorium concentrations in the Darling Scarp area to the east of the city of Perth in Western Australia. It was considered desirable to measure the concentrations of the 220 Rn (thoron) in houses built near such areas and compare these measurements with similar measurements at other sites within Australia. A monitor developed for surveying 220 Rn and 222 Rn levels in Japan was considered most suitable for this purpose. The monitor used for the survey was a passive radon-thoron discriminative dosimeter, similar in type to that used to measure thoron concentrations in traditional Japanese houses. Houses for this study were selected from areas in the Darling Scarp. To provide data on what might be considered typical thoron concentrations associated with other soil types within Australia, a small number of houses in areas on the coastal plain of Western Australia and from the coastal areas of Victoria were also monitored. The mean thoron concentration in homes on the Darling Scarp of 14.0 Bq.m -3 was higher than that in the coastal plain in Western Australia (3.0 Bq.m -3 ) but comparable to the measurements in Victoria (11.3 Bq.m -3 ). For the radon measurements in homes, the mean Scarp concentration was typically 49.2 Bq.m -3 while the WA coastal plain and in Victoria mean concentrations were 11.6 Bq.m -3 and 15.4 Bq.m -3 respectively. For outdoors in the Darling Scarp, the thoron mean was 76.3 Bq.m -3 while the radon level was 18.2 Bq.m -3 . These measurements would suggest that the source of the thoron in homes on the Scarp is from the underlying soil rather than from the building materials. (author)

  15. Diffusion measurement in ferrous infused gel dosimeters

    International Nuclear Information System (INIS)

    Zahmatkesh, M. H.; Healy, B. J.

    2003-01-01

    Background: The compositions of Ferrous sulphate, Agarose and Xylenol orange dye and Ferrous sulphate, Gelatin and Xylenol orange dye in solution of distilled water and sulphuric acid are two tissue-equivalent gel dosimeters. Ionizing radiation causes oxidation of Fe 2+ ion to Fe 3+ ions which diffuse through the gel matrix and blur the image of absorbed dose over a period of hours after irradiation. Materials and methods: 25 m M sulphuric acid, 0.4 mm ferrous ammonium sulphate, 0.2 mm xylenol orange dye and 1% by weight agarose in distilled water named Agarose and Xylenol orange dye and 0.1 mm ferrous ammonium sulphate, 0.1 mm xylenol orange dye, 50 mm sulphuric acid and 5% by weight gelatin in distilled water named Gelatin and Xylenol orange dye are used as two gel dosimeters. All chemicals were supplied by Sigma Ald ridge Company, Germany. The gels were poured in Perspex casts and were irradiated to a beam of X ray from linear accelerators or X ray machine. Results: In this study diffusion coefficients of Agarose and Xylenol orange dye and Gelatin and Xylenol orange dye dosimeters have been measured through a computer program for different temperature. The ferric ion diffusion coefficient (D) for the Agarose and Xylenol orange dye and Gelatin and Xylenol orange dye dosimeters were measured as (1.19.±0.03) x 10 -2 cm 2 .hr -1 and (0.83±0.03) x 10 -2 cm 2 .hr -1 respectively at room temperature. Conclusion: For both dosimeters the diffusion coefficients decreased with gel storage temperatures down to 6 d ig C . Gelatin and Xylenol orange dye dosimeters have advantage of lower diffusion coefficient for a specified temperature

  16. Site-specific calibration of the Hanford personnel neutron dosimeter

    International Nuclear Information System (INIS)

    Endres, A.W.; Brackenbush, L.W.; Baumgartner, W.V.; Rathbone, B.A.

    1994-10-01

    A new personnel dosimetry system, employing a standard Hanford thermoluminescent dosimeter (TLD) and a combination dosimeter with both CR-39 nuclear track and TLD-albedo elements, is being implemented at Hanford. Measurements were made in workplace environments in order to verify the accuracy of the system and establish site-specific factors to account for the differences in dosimeter response between the workplace and calibration laboratory. Neutron measurements were performed using sources at Hanford's Plutonium Finishing Plant under high-scatter conditions to calibrate the new neutron dosimeter design to site-specific neutron spectra. The dosimeter was also calibrated using bare and moderated 252 Cf sources under low-scatter conditions available in the Hanford Calibration Laboratory. Dose equivalent rates in the workplace were calculated from spectrometer measurements using tissue equivalent proportional counter (TEPC) and multisphere spectrometers. The accuracy of the spectrometers was verified by measurements on neutron sources with calibrations directly traceable to the National Institute of Standards and Technology (NIST)

  17. Development of a new type thyroid glands dosimeter

    International Nuclear Information System (INIS)

    He Lihua; Song Yiyang; Chen Qin; Chen Yannan

    2000-01-01

    A new dosimeter of 125 I in thyroid gland is described. The dosimeter consists of NaI(Tl) detector and intelligent data recorder. Single-chip-microcomputer is used for data handling. The activity of 125 I in thyroid glands of human being is measured directly, rapidly, and accurately. Furthermore, it can calculate and display the intake, committed dose equivalent and committed effective dose equivalent. The measuring range of 125 I in thyroid glands is 10-2 x 10 6 Bq. The dosimeter has been operating continuously for a long time with high stability

  18. Device for the automatic evaluation of pencil dosimeters

    International Nuclear Information System (INIS)

    Schallopp, B.

    1976-01-01

    In connenction with the automation of radiation protection in nuclear power plants, an automatic reading device has been developed for the direct input of the readings of pencil dosimeters into a computer. Voltage measurements would be simple but are excluded, because the internal electrode of the dosimeter may not be touched, for operational reasons. This paper describes an optical/electronic conversion device in which the reading of the dosimeter is projected onto a Vidicon, scanned, and converted into a digital signal for output to the computer. (orig.) [de

  19. Portable battery-free charger for radiation dosimeters

    International Nuclear Information System (INIS)

    Manning, F.W.

    1984-01-01

    This invention is a novel portable charger for dosimeters of the electrometer type. The charger does not require batteries or piezoelectric crystals and is of rugged construction. In a preferred embodiment, the charge includes a housing which carries means for mounting a dosimeter to be charged. The housing also includes contact means for impressing a charging voltage across the mounted dosimeter. Also, the housing carries a trigger for operating a charging system mounted in the housing. The charging system includes a magnetic loop including a permanent magnet for establishing a magnetic field through the loop. A segment of the loop is coupled to the trigger for movement thereby to positions opening and closing the loop. A coil inductively coupled with the loop generates coil-generated voltage pulses when the trigger is operated to open and close the loop. The charging system includes an electrical circuit for impressing voltage pulses from the coil across a capacitor for integrating the pulses and applying the resulting integrated voltage across the above-mentioned contact means for charging the dosimeter

  20. Portable battery-free charger for radiation dosimeters

    Science.gov (United States)

    Manning, Frank W.

    1984-01-01

    This invention is a novel portable charger for dosimeters of the electrometer type. The charger does not require batteries or piezoelectric crystals and is of rugged construction. In a preferred embodiment, the charge includes a housing which carries means for mounting a dosimeter to be charged. The housing also includes contact means for impressing a charging voltage across the mounted dosimeter. Also, the housing carries a trigger for operating a charging system mounted in the housing. The charging system includes a magnetic loop including a permanent magnet for establishing a magnetic field through the loop. A segment of the loop is coupled to the trigger for movement thereby to positions opening and closing the loop. A coil inductively coupled with the loop generates coil-generated voltage pulses when the trigger is operated to open and close the loop. The charging system includes an electrical circuit for impressing voltage pulses from the coil across a capacitor for integrating the pulses and applying the resulting integrated voltage across the above-mentioned contact means for charging the dosimeter.

  1. Water-equivalent one-dimensional scintillating fiber-optic dosimeter for measuring therapeutic photon beam

    International Nuclear Information System (INIS)

    Moon, Jinsoo; Won Jang, Kyoung; Jae Yoo, Wook; Han, Ki-Tek; Park, Jang-Yeon; Lee, Bongsoo

    2012-01-01

    In this study, we fabricated a one-dimensional scintillating fiber-optic dosimeter, which consists of 9 scintillating fiber-optic dosimeters, septa, and PMMA blocks for measuring surface and percentage depth doses of a therapeutic photon beam. Each dosimeter embedded in the 1-D scintillating fiber-optic dosimeter is composed of square type organic scintillators and plastic optical fibers. Also black PVC films are used as septa to minimize cross-talk between the scintillating fiber-optic dosimeters. To construct a dosimeter system, a 1-D scintillating fiber-optic dosimeter and a CMOS image sensor were combined with 20 m-length plastic optical fibers. Using the dosimeter system, we measured surface and percentage depth doses of 6 and 15 MV photon beams and compared the results with those of EBT films and an ionization chamber. - Highlights: ► Fabrication of a one-dimensional scintillating fiber-optic dosimeter. ► The one-dimensional scintillating fiber-optic dosimeter has 9 scintillating fiber-optic dosimeters. ► Measurements of surface and percentage depth doses of a therapeutic photon beam. ► The results were compared with those of EBT films and an ionization chamber.

  2. Development of direct reading dosimeters for the dose 0-3 mSv and 0-5 mSv ranges for personnel monitoring

    International Nuclear Information System (INIS)

    Gaikwad, P.V.; Shirkar, Y.B.; Patil, A.S.; Madgaonkar, P.P.; Kale, K.L.; Guhagarkar, H.V.; Gandhi, D.P.; Gupta, S.K.; Kothiyal, G.P.; Sahni, V.C.

    1998-01-01

    Direct reading dosimeters (DRDs) are widely used to measure cumulative dose received by personnel working at nuclear reactor sites or in other environment having x- and gamma rays. A DRD operates on the principle of gold leaf electroscope, and is a small, rugged, hermetically sealed, self reading type device easily carried by an individual in his pocket. The development of dosimeters suitable for the dose ranges 0-3 mSv and 0-5 mSv is reported

  3. Fast fluence measurement for JOYO irradiation field using niobium dosimeter

    International Nuclear Information System (INIS)

    Ito, Chikara

    2004-03-01

    Neutron fluence and spectrum are key parameters in various irradiation tests and material surveillance tests so they need to be evaluated accurately. The reactor dosimetry test has been conducted by the multiple foil activation method, and a niobium dosimeter has been developed for measurement of fast neutron fluence in the experimental fast reactor JOYO. The inelastic scattering reaction of 93 Nb has a low threshold energy, about 30 keV, and the energy distribution of reaction cross section is similar to the displacement cross section for iron. Therefore, a niobium dosimeter is suitable for evaluation of the fast neutron fluence and the displacement per atom for iron. Moreover, a niobium dosimeter is suited to measure neutron fluence in long-term irradiation test because 93 Nb, which is produced by the reaction, has a long half-life (16.4 years). This study established a high precision measurement technique using the niobium reaction rate. The effect of self-absorption was decreased by the solution and evaporation to dryness of niobium dosimeter. The dosimeter weight was precisely measured using the inductively coupled plasma mass spectrometer. This technique was applied to JOYO dosimetry. The fast neutron fluences (E > 0.1 MeV) found by measuring the reaction rate in the niobium dosimeter were compared with the values evaluated using the multiple foil activation method. The ratio of measured fast neutron fluences by means of niobium dosimeter and multiple foil activation method range from 0.97 to 1.03 and agree within the experimental uncertainty. The measurement errors of fast neutron fluence by niobium dosimeter range from 4.5% (fuel region) to 10.1% (in-vessel storage rack). As a result of this study, the high precision measurement of fast neutron fluence by niobium dosimeters was confirmed. The accuracy of fast reactor dosimetry will be improved by application of niobium dosimeters to the irradiation tests in the JOYO MK-III core. (author)

  4. Assessment of radiation exposure of nuclear medicine staff using personal TLD dosimeters and charcoal detectors

    Energy Technology Data Exchange (ETDEWEB)

    Jimenez, F.; Garcia-Talavera, M.; Pardo, R.; Deban, L. [Valladolid Univ., Dept. de Quimica Analitica, Facultad de Ciencias (Spain); Garcia-Talavera, P.; Singi, G.M.; Martin, E. [Hospital Clinico Univ., Servicio de Medicina Nuclear, Salamanca (Spain)

    2006-07-01

    Although the main concern regarding exposure to ionizing radiation for nuclear medicine workers is external radiation, inhalation of radionuclides can significantly contribute to the imparted doses. We propose a new approach to assess exposure to inhalation of {sup 131}I based on passive monitoring using activated charcoal detectors. We compared the inhalation doses to the staff of a nuclear medicine department, based on the measurements derived from charcoal detectors placed at various locations, and the external doses monitored using personal TLD dosimeters. (authors)

  5. Assessment of radiation exposure of nuclear medicine staff using personal TLD dosimeters and charcoal detectors

    International Nuclear Information System (INIS)

    Jimenez, F.; Garcia-Talavera, M.; Pardo, R.; Deban, L.; Garcia-Talavera, P.; Singi, G.M.; Martin, E.

    2006-01-01

    Although the main concern regarding exposure to ionizing radiation for nuclear medicine workers is external radiation, inhalation of radionuclides can significantly contribute to the imparted doses. We propose a new approach to assess exposure to inhalation of 131 I based on passive monitoring using activated charcoal detectors. We compared the inhalation doses to the staff of a nuclear medicine department, based on the measurements derived from charcoal detectors placed at various locations, and the external doses monitored using personal TLD dosimeters. (authors)

  6. Laser readable thermoluminescent radiation dosimeters and methods for producing thereof

    International Nuclear Information System (INIS)

    Braunlich, P.F.; Tetzlaff, W.

    1989-01-01

    Thin layer thermoluminescent radiation dosimeters for use in laser readable dosimetry systems, and methods of fabricating such thin layer dosimeters are disclosed. The thin layer thermoluminescent radiation dosimeters include a thin substrate made from glass or other inorganic materials capable of withstanding high temperatures and high heating rates. A thin layer of a thermoluminescent phosphor material is heat bonded to the substrate using an inorganic binder such as glass. The dosimeters can be mounted in frames and cases for ease in handling. Methods of the invention include mixing a suitable phosphor composition and binder, both being in particulate or granular form. The mixture is then deposited onto a substrate such as by using mask printing techniques. The dosimeters are thereafter heated to fuse and bond the binder and phosphor to the substrate. 34 figs

  7. Energy response study of modified CR-39 neutron personnel dosimeter

    International Nuclear Information System (INIS)

    Sathian, Deepa; Bakshi, A.K.; Datta, D.; Nair, Sreejith S.; Sathian, V.; Mishra, Jitendra; Sen, Meghnath

    2018-01-01

    Personnel neutron dosimetry is an integral part of radiation protection. No single dosimeter provides the satisfactory energy response, sensitivity, angular dependence characteristics and accuracy necessary to meet the requirement of an ideal personnel neutron dosimeter. The response of a personnel neutron dosimeter is critically dependent upon the energy distribution of the neutron field. CR-39 personnel neutron dosimeters were typically calibrated in the standard neutron field of 252 Cf and 241 Am-Be in our laboratory, although actual neutron fields may vary from the calibration neutron spectrum. Recently the badge cassette of the personnel neutron dosimeter was changed due to frequent damage of the PVC badge used earlier. This paper discusses energy response of CR-39 solid state nuclear track detector loaded in this modified badge cassette as per latest ISO recommendation

  8. Feasibility of Ultra-Thin Fiber-Optic Dosimeters for Radiotherapy Dosimetry.

    Science.gov (United States)

    Lee, Bongsoo; Kwon, Guwon; Shin, Sang Hun; Kim, Jaeseok; Yoo, Wook Jae; Ji, Young Hoon; Jang, Kyoung Won

    2015-11-17

    In this study, prototype ultra-thin fiber-optic dosimeters were fabricated using organic scintillators, wavelength shifting fibers, and plastic optical fibers. The sensor probes of the ultra-thin fiber-optic dosimeters consisted of very thin organic scintillators with thicknesses of 100, 150 and 200 μm. These types of sensors cannot only be used to measure skin or surface doses but also provide depth dose measurements with high spatial resolution. With the ultra-thin fiber-optic dosimeters, surface doses for gamma rays generated from a Co-60 therapy machine were measured. Additionally, percentage depth doses in the build-up regions were obtained by using the ultra-thin fiber-optic dosimeters, and the results were compared with those of external beam therapy films and a conventional fiber-optic dosimeter.

  9. Dosimetric properties of the pocket alarm dosimeter type Graetz TDW 10/20

    International Nuclear Information System (INIS)

    Hauser, M.; Burgkhardt, B.; Piesch, E.

    1980-08-01

    In personnel monitoring pocket dosimeters with build-in alarm devices are increasingly in use. The report presents results of a test performed at Karlsruhe. The properties investigated are above all linearity and reproducibility of the dose reading as well as of the acustic alarm indication, dependence on the dose reading, the photon energy, the direction of the radiation incidence, the dose rate, the temperature and operational characteristic of the batteries. (orig./HP) [de

  10. UVB DNA dosimeters analyzed by polymerase chain reactors

    International Nuclear Information System (INIS)

    Yoshida, Hiroko; Regan, J.D.; Florida Inst. of Tech., Melbourne, FL

    1997-01-01

    Purified bacteriophage λ DNA was dried on a UV-transparent polymer film and served as a UVB dosimeter for personal and ecological applications. Bacteriophage λ DNA was chosen because it is commercially available and inexpensive, and its entire sequence is known. Each dosimeter contained two sets of DNA sandwiched between UV-transparent polymer films, one exposed to solar radiation (experimental) and another protected from UV radiation by black paper (control). The DNA dosimeter was then analyzed by a polymerase chain reaction (PCR) that amplifies a 500 base pair specific region of λ DNA. Photoinduced damage in DNA blocks polymerase from synthesizing a new strand; therefore, the amount of amplified product in UV-exposed DNA was reduced from that found in control DNA. The dried λ DNA dosimeter is compact, robust, safe and transportable, stable over long storage times and provides the total UVB dose integrated over the exposure time. (author)

  11. A pocket warning γ-dosimeter with numerical display

    International Nuclear Information System (INIS)

    Jones, A.R.

    1980-09-01

    A pocket warning dosimeter is described. It provides alarms (continuous tone and a flashing red light) when a presettable dose has been accumulated in the range .064 - 16.4 rads (0.64 - 164 μGy). This warning level can be selected in nine steps of 2 with a switch inside the dosimeter. The dose rate is indicated by a series of sound pulses whose repetition rate is proportional to the dose rate. At 1 rad/h (10 mGy/h) about 17 pluses/minute are emitted. The accumulated dose up to 20 rads (0.2 Gy) is displayed in steps of 1 mrad (10 μGy) with a liquid crystal display. A red LED lights before battery failure occurs. The effects of changes in temperature, battery voltage, dose rate and photon energy upon dosimeter sensitivity are presented. Finally, the applications of the dosimeter are discussed. (auth)

  12. LOW-COST PERSONNEL DOSIMETER.

    Science.gov (United States)

    specification was achieved by simplifying and improving the basic Bendix dosimeter design, using plastics for component parts, minimizing direct labor, and making the instrument suitable for automated processing and assembly. (Author)

  13. Dependence of alanine gel dosimeter response as a function of photon clinical beams dose rate

    International Nuclear Information System (INIS)

    Silva, Cleber Feijo; Campos, Leticia Lucente

    2013-01-01

    Gel dosimetry is a new area developed by Gore, it is ery useful for application in radiotherapy because using NMR imaging as evaluation technique is possible to evaluate three dimensional absorbed dose distribution. The measure technique is based on difference of ferrous (Fe 2+ ) and ferric (Fe 3+ ) ) ions concentration that can be measured also by spectrophotometry technique. The Alanine gel dosimeter was developed at IPEN. The alanine is an amino acid and tissue equivalent material that presents significant improvement on previous alanine dosimetry systems. The addition of Alanine increases the production of ferric ions in the solution. This work aims to study the dose rate dependence of photon clinical beams radiation on the alanine gel dosimeter optical response, as well as the response repeatability and gel production reproducibility, since this property is very important for characterization and standardization of any dosimeter. (author)

  14. Angular dependence of the nanoDot OSL dosimeter

    International Nuclear Information System (INIS)

    Kerns, James R.; Kry, Stephen F.; Sahoo, Narayan; Followill, David S.; Ibbott, Geoffrey S.

    2011-01-01

    Purpose: Optically stimulated luminescent detectors (OSLDs) are quickly gaining popularity as passive dosimeters, with applications in medicine for linac output calibration verification, brachytherapy source verification, treatment plan quality assurance, and clinical dose measurements. With such wide applications, these dosimeters must be characterized for numerous factors affecting their response. The most abundant commercial OSLD is the InLight/OSL system from Landauer, Inc. The purpose of this study was to examine the angular dependence of the nanoDot dosimeter, which is part of the InLight system. Methods: Relative dosimeter response data were taken at several angles in 6 and 18 MV photon beams, as well as a clinical proton beam. These measurements were done within a phantom at a depth beyond the build-up region. To verify the observed angular dependence, additional measurements were conducted as well as Monte Carlo simulations in MCNPX. Results: When irradiated with the incident photon beams parallel to the plane of the dosimeter, the nanoDot response was 4% lower at 6 MV and 3% lower at 18 MV than the response when irradiated with the incident beam normal to the plane of the dosimeter. Monte Carlo simulations at 6 MV showed similar results to the experimental values. Examination of the results in Monte Carlo suggests the cause as partial volume irradiation. In a clinical proton beam, no angular dependence was found. Conclusions: A nontrivial angular response of this OSLD was observed in photon beams. This factor may need to be accounted for when evaluating doses from photon beams incident from a variety of directions.

  15. Polymer gel dosimeter with AQUAJOINT® as hydrogel matrix

    Science.gov (United States)

    Maeyama, Takuya; Ishida, Yasuhiro; Kudo, Yoshihiro; Fukasaku, Kazuaki; Ishikawa, Kenichi L.; Fukunishi, Nobuhisa

    2018-05-01

    We report a polymer gel dosimeter based on a new gel matrix (AQUAJOINT®) that is a thermo-irreversible hydrogel formed by mixing two types of water-based liquids at room temperature. Normoxic N-vinylpyrrolidone-based polymer gels were prepared with AQUAJOINT® instead of gelatin. This AQUAJOINT®-based gel dosimeter exhibits a 2.5-fold increase in sensitivity over a gelatin-based gel dosimeter and a linear dose-response in the dose range of 0-8 Gy. This gel has heat resistance in a jar and controlled gel properties such as viscoelastic and mechanical characters, which may be useful for deformable polymer gel dosimetry.

  16. Alanine EPR dosimeter response in proton therapy beams

    International Nuclear Information System (INIS)

    Gall, K.; Serago, C.; Desrosiers, M.; Bensen, D.

    1997-01-01

    We report a series of measurements directed to assess the suitability of alanine as a mailable dosimeter for dosimetry quality assurance of proton radiation therapy beams. These measurements include dose-response of alanine at 140 MeV, and comparison of response vs energy with a parallel plate ionization chamber. All irradiations were made at the Harvard Cyclotron Laboratory, and the dosimeters were read at NIST. The results encourage us that alanine could be expected to serve as a mailable dosimeter with systematic error due to differential energy response no greater than 3% when doses of 25 Gy are used. (Author)

  17. Automated dose estimation for lost or damaged dosimeters

    International Nuclear Information System (INIS)

    Thompson, W.L.; Deininger, R.J.

    1988-01-01

    This paper reports that some dosimetry vendors will compute doses for their customers' lost/damaged dosimeters based upon an average of recent dosimeter readings. However, the vendors usually require authorization from the customer for each such occurrence. Therefore, the tedious task of keeping track of the overdue status of each missing dosimeter and constantly notifying the vendor is still present. Also, depending on the monthly variability of a given person's doses, it may be more valid to use the employee's average dose, his/her highest dose over a recent period, an average dose of other employees with similar job duties for that period, or the maximum permissible dose. Thus, the task of estimating doses for lost/damaged dosimeters cannot be delegated to dosimetry vendor. Instead, the radiation safety department must sue the data supplied by the vendor as input for performing estimates. The process is performed automatically at the Medical Center Hospital of Vermont using a personal computer and a relational database

  18. Investigating potential physicochemical errors in polymer gel dosimeters

    International Nuclear Information System (INIS)

    Sedaghat, Mahbod; Lepage, Martin; Bujold, Rachel

    2011-01-01

    Measurement errors in polymer gel dosimetry can originate either during irradiation or scanning. One concern related to the exothermic nature of polymerization reaction was that the heat released in polymer gel dosimeters during irradiation modifies their dose response. In this paper, the effect of heat released from the exothermal polymerization reaction on the dose response of a number of dosimeters was studied. In addition, we investigated whether heat-generated geometric distortion existed in newly proposed gel dosimeters that contain highly thermoresponsive polymers. Our results suggest that despite a significant internal temperature increase in some gel compositions, their dose responses are not affected when oxygen is well expelled mechanically from the gel mixture. We also report on significant pre-irradiation instability in some recently developed polymer gel dosimeters but that geometric distortions were not observed. Data obtained by a set of small calibration vials are compared to those obtained from larger phantoms, and potential physicochemical causes of deviations between them are identified.

  19. Investigating potential physicochemical errors in polymer gel dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Sedaghat, Mahbod; Lepage, Martin [Centre d' imagerie moleculaire de Sherbrooke, Departement de medecine nucleaire et radiobiologie, Universite de Sherbrooke, Sherbrooke, QC (Canada); Bujold, Rachel, E-mail: martin.lepage@usherbrooke.ca [Service de radio-oncologie, Centre hospitalier universitaire de Sherbrooke, Sherbrooke, QC (Canada)

    2011-09-21

    Measurement errors in polymer gel dosimetry can originate either during irradiation or scanning. One concern related to the exothermic nature of polymerization reaction was that the heat released in polymer gel dosimeters during irradiation modifies their dose response. In this paper, the effect of heat released from the exothermal polymerization reaction on the dose response of a number of dosimeters was studied. In addition, we investigated whether heat-generated geometric distortion existed in newly proposed gel dosimeters that contain highly thermoresponsive polymers. Our results suggest that despite a significant internal temperature increase in some gel compositions, their dose responses are not affected when oxygen is well expelled mechanically from the gel mixture. We also report on significant pre-irradiation instability in some recently developed polymer gel dosimeters but that geometric distortions were not observed. Data obtained by a set of small calibration vials are compared to those obtained from larger phantoms, and potential physicochemical causes of deviations between them are identified.

  20. Validation of an Innovative Satellite-Based UV Dosimeter

    Science.gov (United States)

    Morelli, Marco; Masini, Andrea; Simeone, Emilio; Khazova, Marina

    2016-08-01

    We present an innovative satellite-based UV (ultraviolet) radiation dosimeter with a mobile app interface that has been validated by exploiting both ground-based measurements and an in-vivo assessment of the erythemal effects on some volunteers having a controlled exposure to solar radiation.Both validations showed that the satellite-based UV dosimeter has a good accuracy and reliability needed for health-related applications.The app with this satellite-based UV dosimeter also includes other related functionalities such as the provision of safe sun exposure time updated in real-time and end exposure visual/sound alert. This app will be launched on the global market by siHealth Ltd in May 2016 under the name of "HappySun" and available both for Android and for iOS devices (more info on http://www.happysun.co.uk).Extensive R&D activities are on-going for further improvement of the satellite-based UV dosimeter's accuracy.

  1. Performance testing of beta dosimeters used at Department of Energy facilities

    International Nuclear Information System (INIS)

    Roberson, P.L.; Holbrook, K.L.; Pappin, J.L.

    1983-01-01

    A performance test based on the American national draft standard N13.11 was conducted for dosimeter systems in use at Department of Energy facilities. The large differences in dosimeter response found were due to use of different calibration source standards and different dosimeter designs. Differences in 90 Sr/ 90 Y calibrations were approximately 20% or less for all but one participant. The differences observed were attributed to variable thicknesses of dosimeter elements and variable source irradiation geometries. Improved beta calibration standards will result if irradiation specifications include acceptable ranges from the depth-dose characteristics. The low-energy beta responses observed were consistent with the thicknesses of dosimeter sensitive elements and overlying filtration

  2. Using a thermoluminescent dosimeter to evaluate the location reliability of the highest–skin dose area detected by treatment planning in radiotherapy for breast cancer

    Energy Technology Data Exchange (ETDEWEB)

    Sun, Li-Min, E-mail: limin.sun@yahoo.com [Department of Radiation Oncology, Zuoying Branch of Kaohsiung Armed Forces General Hospital, Kaohsiung City, Taiwan (China); Huang, Chih-Jen [Department of Radiation Oncology, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung City, Taiwan (China); Faculty of Medicine, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung City, Taiwan (China); College of Medicine, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung City, Taiwan (China); Chen, Hsiao-Yun [Department of Radiation Oncology, Kaohsiung Medical University Hospital, Kaohsiung Medical University, Kaohsiung City, Taiwan (China); Meng, Fan-Yun [Department of General Surgery, Zuoying Branch of Kaohsiung Armed Forces General Hospital, Kaohsiung City, Taiwan (China); Lu, Tsung-Hsien [Department of Radiation Oncology, Zuoying Branch of Kaohsiung Armed Forces General Hospital, Kaohsiung City, Taiwan (China); Tsao, Min-Jen [Department of General Surgery, Zuoying Branch of Kaohsiung Armed Forces General Hospital, Kaohsiung City, Taiwan (China)

    2014-01-01

    Acute skin reaction during adjuvant radiotherapy for breast cancer is an inevitable process, and its severity is related to the skin dose. A high–skin dose area can be speculated based on the isodose distribution shown on a treatment planning. To determine whether treatment planning can reflect high–skin dose location, 80 patients were collected and their skin doses in different areas were measured using a thermoluminescent dosimeter to locate the highest–skin dose area in each patient. We determined whether the skin dose is consistent with the highest-dose area estimated by the treatment planning of the same patient. The χ{sup 2} and Fisher exact tests revealed that these 2 methods yielded more consistent results when the highest-dose spots were located in the axillary and breast areas but not in the inframammary area. We suggest that skin doses shown on the treatment planning might be a reliable and simple alternative method for estimating the highest skin doses in some areas.

  3. Using a thermoluminescent dosimeter to evaluate the location reliability of the highest–skin dose area detected by treatment planning in radiotherapy for breast cancer

    International Nuclear Information System (INIS)

    Sun, Li-Min; Huang, Chih-Jen; Chen, Hsiao-Yun; Meng, Fan-Yun; Lu, Tsung-Hsien; Tsao, Min-Jen

    2014-01-01

    Acute skin reaction during adjuvant radiotherapy for breast cancer is an inevitable process, and its severity is related to the skin dose. A high–skin dose area can be speculated based on the isodose distribution shown on a treatment planning. To determine whether treatment planning can reflect high–skin dose location, 80 patients were collected and their skin doses in different areas were measured using a thermoluminescent dosimeter to locate the highest–skin dose area in each patient. We determined whether the skin dose is consistent with the highest-dose area estimated by the treatment planning of the same patient. The χ 2 and Fisher exact tests revealed that these 2 methods yielded more consistent results when the highest-dose spots were located in the axillary and breast areas but not in the inframammary area. We suggest that skin doses shown on the treatment planning might be a reliable and simple alternative method for estimating the highest skin doses in some areas

  4. Approving of personal dosimeter services

    International Nuclear Information System (INIS)

    Bergman, K.; Malmqvist, L.

    2001-09-01

    The Swedish regulation SSI FS 98:5 requires that radiological workers of category A use dosemeters from an approved personal dosimetry service. The regulation also includes certain specific dosimeter requirements, which are based on those presented in the Technical Recommendations by the European Commission (Report EUR 14852 EN, 1994). All services have been tested for their ability to determine Hp(10) and some of them to determine Hp(0.07) at one radiation quality. The test was performed in the interval 0.2 mSv to 100 mSv at three different dose equivalents unknown to the system owner. The 11 services operating in Sweden at the moment use 5 different types of dosimeters. The five unique systems have been tested regarding the angular and energy dependence of the response of the dosimeters. The dosimeters were irradiated to a personal dose equivalent of about 1 mSv at three photon energies and at four angles (0, 20, 40 and 60 deg. resp. ) both vertically and horizontally rotated. Only 2 of the services determine Hp(0.07) for beta and gamma radiation and were tested for this quantity. The test results for Hp(10) are all except two within the trumpet curve. For the unique systems it is shown that the uncertainty related to angular response at a specified energy is within the required ±40 % except for the lowest X-ray quality at 40 kV. The response is more dependent on photon energy than on the direction of the photon radiation and the choice of radiation quality for the calibration is of great importance for the system performance

  5. Water equivalence of polymer gel dosimeters

    International Nuclear Information System (INIS)

    Sellakumar, P.; James Jebaseelan Samuel, E.; Supe, Sanjay S.

    2007-01-01

    To evaluate the water equivalence and radiation transport properties of polymer gel dosimeters over the wide range of photon and electron energies 14 different types of polymer gels were considered. Their water equivalence was evaluated in terms of effective atomic number (Z eff ), electron density (ρ e ), photon mass attenuation coefficient (μ/ρ), photon mass energy absorption coefficient (μ en /ρ) and total stopping power (S/ρ) tot of electrons using the XCOM and the ESTAR database. The study showed that the effective atomic number of polymer gels were very close ( en /ρ for all polymer gels were in close agreement ( tot of electrons in polymer gel dosimeters were within 1% agreement with that of water. From the study we conclude that at lower energy (<80keV) the polymer gel dosimeters cannot be considered water equivalent and study has to be carried out before using the polymer gel for clinical application

  6. Comparison of Panasonic’s Dosimetric System with Gamma-31 Dosimeters

    Directory of Open Access Journals (Sweden)

    Paweł Urban

    2013-01-01

    Full Text Available Equipment being used in medical or industrial institutions is often a source of ionizing radiation with different energies and types, which complicates the detection and assessment of doses. Up until now, for dosimetric measurements of ionizing radiation, Gamma-31 dosimeters have been used in the Central Mining Institute for many years. Now, this system will be expanded by a Panasonic system, for which measurement procedures were developed and comparisons with other dosimeters were held. The method is based on a four-element dosimeters UD-802 Panasonic equipped with CaSO and LiBO detectors additionally sheltered by filters of different surface mass. The use of UD-802 dosimeters, in contrast to Gamma-31 dosimeters, permits measuring radiation doses in a different range of photon energy. Consequently, it is possible to obtain a more accurate analysis of the hazards caused by gamma radiation in underground mines. The publication includes a description of the dosimetry system and presents the results of measurements conducted by means of both types of dosimeters. In order to verify the correctness of the indications of the new dosimetry system a series of measurements were carried out, which allowed examining the behaviour of the dosimeters under different environmental conditions. As a place of exposure, the selected laboratories in the Silesian Centre for Environmental Radiometry were chosen, where the work is connected with (TENORM and equipment producing ionizing radiation or containing sources of this type of radiation. Moreover, to observe the dosimeters behaviour in difficult environmental conditions, they were exposed in water treatment plants and an underground potassium salt mine.

  7. Characterization of Thymol blue Radiochromic dosimeters for high dose applications

    Directory of Open Access Journals (Sweden)

    Feras M. Aldweri

    2018-03-01

    Full Text Available Thymol blue (TB solutions and Thymol blue Polyvinyl Alcohol (TB-PVA films have been introduced as Radiochromic dosimeter for high dose applications. The dosimeters were irradiated with gamma ray (60Co source from 5 to 30 kGy for film, and from 0.150 kGy to 4 kGy for solution. The optical density of unirradiated and irradiated TB solution as well as TB-PVA film dosimeters were studied in terms of absorbance at 434 nm using UV/VIS spectrophotometer. The effects of scan temperature, light pre-gamma irradiation, dose rate, relative humidity and stability of the absorbance of solutions and films after irradiation were investigated. We found the dose sensitivity of TB solution and TB-PVA film dosimeters increases significantly with increases of the absorbed dose as well as with the increases of TB dye concentrations. The useful dose range of developed TB solutions and TB-PVA films dosimeters is in the range 0.125–1 kGy and of 5–20 kGy, respectively. Keywords: Dose sensitivity, Radio-chromic dosimeter, Thymol blue, Absorbance, Concentrations

  8. A simultaneous electron energy and dosimeter calibration method for an electron beam irradiator

    International Nuclear Information System (INIS)

    Tanaka, R.; Sunaga, H.; Kojima, T.

    1991-01-01

    In radiation processing using electron accelerators, the reproducibility of absorbed dose in the product depends not only on the variation of beam current and conveyor speed, but also on variations of other accelerator parameters. This requires routine monitoring of the beam current and the scan width, and also requires periodical calibration of routine dosimeters usually in the shape of film, electron energy, and other radiation field parameters. The electron energy calibration is important especially for food processing. The dose calibration method using partial absorption calorimeters provides only information about absorbed dose. Measurement of average electron current density provides basic information about the radiation field formed by the beam scanning and scattering at the beam window, though it does not allow direct dose calibration. The total absorption calorimeter with a thick absorber allows dose and dosimeter calibration, if the depth profile of relative dose in a reference absorber is given experimentally. It also allows accurate calibration of the average electron energy at the surface of the calorimeter core, if electron fluence received by the calorimeter is measured at the same time. This means that both electron energy and dosimeters can be simultaneously calibrated by irradiation of a combined system including the calorimeter, the detector of the electron current density meter, and a thick reference absorber for depth profile measurement of relative dose. We have developed a simple and multifunctional system using the combined calibration method for 5 MeV electron beams. The paper describes a simultaneous calibration method for electron energy and film dosimeters, and describes the electron current density meter, the total absorption calorimeter, and the characteristics of this method. (author). 13 refs, 7 figs, 3 tabs

  9. Indoor radon measurement in some adobe houses in the Kassena Nankana area of the Upper East Region

    International Nuclear Information System (INIS)

    Quashie, F. K.

    2010-06-01

    Inhalation of radon and its daughter products is the major contributor to the total exposure of the population to natural radiation. The present study has measured radon gas concentration in some Adobe houses and the soil radon gas around these houses in the Kassena Nankana Area of the Upper East Region by using passive radon indoor dosimeter containing solid-state nuclear track detector (SSNTD) commercially known as LR - 115 (type II, pelliculable). Fifty (50) indoor radon dosimeters were placed in the various Adobe houses in the study area. Additionally, thirty (30) dosimeters were placed in the soil around some of the houses at a depth of 75 cm. Soil radon dosimeters were retrieved after two (2) weeks while the indoor radon dosimeters were retrieved after 78 to 82 days. The detectors were then chemically etched. The digital laser optic system and the spark counter coupled with microfiche reader were used in counting both the indoor and soil detectors respectively. Indoor radon concentration in the study area range from 35.28 Bq/m 3 to 244.22 Bq/m 3 . An active dosimeter known as the radon scout plus which gives instantaneous readings between 1 to 3 hours was also used in nineteen of the adobe houses in the study area and a total average radon concentration of 56.90 Bq/m3 was obtained. The soil radon concentration was also found to range from 2.12 kBq/m 3 to 15.03 kBq/m 3 . A good correlation was found to exist between the soil radon concentration and that of the indoor radon concentration with a correlation coefficient of about 0.61. The mean radon emanation coefficient of some fifteen (15) soil samples monitored was 0.46. The average annual effective dose was estimated to be about 1.66 mSv/y and that of the average annual effective dose using the equilibrium equivalent factor (F) was 1.00 mSv/y

  10. Personnel radiation monitoring by thermoluminescence dosimetry (2000-2001)

    International Nuclear Information System (INIS)

    Mi-Cho-Cho, Daw; Hla-Hla-Win, Daw; Thin-Thin-Kraing

    2001-01-01

    Thermoluminescence dosimetry service was introduced in 1991. Personnel who exposed directly or indirectly to radiation is monitored by thermoluminescent dosimeter. TL materials used for thermoluminescent dosimeter are in the form of thin disc. Personnel whole-body and extremity doses are measured by thermoluminescence dosimetry. The Harshaw Model 4500 TLD reader and Vinten 654E TLD reader are used for TLD evaluation. At present about 600 radiation workers are provided with TLD for routine monitoring. It was found that most personnel had received within permissible dose recommended by the International Commission on Radiological Protection (ICRP). (author)

  11. Portable dosimeter

    International Nuclear Information System (INIS)

    Buffa, A.; Caley, R.; Pfaff, K.

    1986-01-01

    A simple but very accurate portable dosimeter is described for indicating the intensity of ionizing radiation, comprising, as a unit: (a) a radiation-detection chamber having a pair of parallel, facing, electrically-conducting, radiation-permeable electrodes spaced from each other to define a volume for a gas which is ionized by the radiation when exposed thereto; (b) electric potential supply means connected across the electrodes for attracting the gas ions to the electrodes and transferring their charge to the electrodes; (c) detection circuit means connected across the electrodes and having at least one of high-frequency electromagnetic- and radiation-sensitive components for detecting the charge on the electrodes and indicating therefrom a representation of the intensity of the radiation; (d) radiation shield means surrounding the radiation-sensitive components of the detection circuit means for shielding the latter from the ionizing radiation; (e) electric shield means surrounding the sensitive components of the detection circuit means for shielding the latter from electromagnetic interference including any caused by the ionizing radiation; and (f) ion shield means potting the ion-sensitive components for shielding them from radiation-caused ambient ionization; whereby the entire dosimeter may be assembled as the unit and portably transported into various radiation sources

  12. Advances in the development of Cr-39 based neutron dosimeters

    International Nuclear Information System (INIS)

    Hadlock, D.E.; Parkhurst, M.A.

    1987-12-01

    A combination thermoluminescent dosimeter (TLD) and track etch dosimeter (TED), which can be used for detecting neutrons over a wide energy range, has been developed through recent research in passive neutron dosimetery. This dosimeter uses Li-600 TLDs to detect thermal and low energy neutrons reflected from the body, and the TED polymer of CR-39, to detect fast neutrons from proton recoil interactions with the polyethylene radiator or with CR-39 itself. Some form of the combination dosimeter is currently in use at several US Department of Energy (DOE) facilities, and its use is expected to expand over the next year to include all DOE facilities where significant neutron exposures may occur. The extensive research conducted on the TED component over the past six years has continually focused on material improvements, reduction in processing time and dosimeter handling, and ease of sample readout with the goal of automating the process as much as possible. 1 fig

  13. Wide-area, real-time monitoring and visualization system

    Science.gov (United States)

    Budhraja, Vikram S.; Dyer, James D.; Martinez Morales, Carlos A.

    2013-03-19

    A real-time performance monitoring system for monitoring an electric power grid. The electric power grid has a plurality of grid portions, each grid portion corresponding to one of a plurality of control areas. The real-time performance monitoring system includes a monitor computer for monitoring at least one of reliability metrics, generation metrics, transmission metrics, suppliers metrics, grid infrastructure security metrics, and markets metrics for the electric power grid. The data for metrics being monitored by the monitor computer are stored in a data base, and a visualization of the metrics is displayed on at least one display computer having a monitor. The at least one display computer in one said control area enables an operator to monitor the grid portion corresponding to a different said control area.

  14. Neutron dosimeters and survey meters in accelerators, reactors and other neutron environments

    International Nuclear Information System (INIS)

    1989-03-01

    Neutron fields in occupationally accessible areas around nuclear reactors, radioisotope sources and medical and high energy accelerators have been characterized using currently available information. Neutron, rem meters, such as the Leake detector, are the most suitable instruments available for conducting neutron dose rate surveys in the vicinity of radioisotope neutron sources, nuclear reactors and medical accelerators. However, these instruments have been shown to be deficient in that they overrespond by a factor of four to neutrons in the 0.1 to 1 MeV range and are insensitive to neutrons from about 1 eV up to about 10 keV. Also, they are insensitive to neutrons above 20 MeV and their use must be restricted near high energy accelerators where significant numbers of neutrons above 20 MeV are known to be present. The most suitable instrument of measure dose from neutrons above 20 MeV is the 12 C(n,2n) 11 C scintillation chamber. Commercially available rem meters frequently use BF 3 counters in the pulse mode to detect thermal neutrons. Therefore, measurements around pulsed accelerators must be made with caution to ensure that the detector is not saturated during each pulse and that the accelerator pulse period is greater than the response time of the detector. The personal neutron dosimeters currently available either are known to be insensitive to neutrons above 20 MeV or have not been tested. Also, all except the albedo dosimeter are insensitive to or have not been tested for neutron energies in the range 1 eV to 10 keV. Several dosimeter types respond reasonably well to neutrons in the energy range 0.1 to 15 MeV, for example, CR-39, bubble and superheated drop detectors. However, the first gas a lower limit of sensitivity of about 0.3 mSv. The bubble detector can be designed to measure doses as small as 1μSv and offers the additional benefit of direct-reading capability. The superheated drop detector is not suitable for use around pulsed accelerators because

  15. Design of calibration method in neutron and individual dosimeter

    International Nuclear Information System (INIS)

    Belkhodia, M.

    1984-12-01

    Usually albedo dosemeters are calibrated with beam of monoenergetic neutrons. Since neutron energy around neutron sources varies greatly, we applied the calibration method to a mixed field whose energy spectrum lies between 0.025 ev and 10 Mev. The method is based on a mathematical model that deals with the dosimeter response as a function at the neutron energy. The measurements carried out with solid state nuclear track detectors show the dosimeter practical aspect. The albedo dosimeter calibration gave results on good agreement with the international institution recommendations

  16. Radiation sensitive polymer gel dosimeters

    International Nuclear Information System (INIS)

    Lepage, M.; Back, S.A.J.; Baldock, C.; Whittaker, A.K.; Rintoul, L.

    2000-01-01

    Full text: Radiation sensitive gels are studied for their potential to retain a permanent 3D dose distribution for applications in radiotherapy. Co-monomers dissolved in a tissue-equivalent hydrogel undergo a polymerization reaction upon absorption of ionizing radiation. The polymer formed influences the local spin-spin relaxation time (T 2 ) of the dosimeter that can be determined using magnetic resonance imaging (MRI). The relationship between T2 and the absorbed dose was studied for different initial chemical compositions. The aim was to find a model linking the changes in T 2 with absorbed dose to the initial composition of the dosimeter. It is believed this will help designing new gel dosimeters having desired properties to minimize the uncertainty in the determination of the dose distribution. 1 H, 13 C nuclear magnetic resonance spectroscopy and FT-Raman spectroscopy were used to quantify the amount of monomers still remaining after the absorption of a given dose of radiation. This data is used to model the changes of T2 as a function of the absorbed dose. A model of fast exchange of magnetization between three proton pools was used, where the fraction of protons (f x H ) in the x th pool is obtained from the chemical composition of the dosimeter and the apparent T2 of each pool is determined for a given composition. Initially, the protons are contained in two pools; a mobile (mob), which contains the water protons and the monomers protons, and a gelatin (gela) proton pool. The mobile pool is partially depleted as polymer is formed, the protons are transferred into the polymer (pol) pool. In the figure, the experimental data along with the calculated values are plotted for three different monomer concentrations, with the gelatin concentration fixed. The model is seen to provide a good fit to the experimental data

  17. The LLNL CR-39 personnel neutron dosimeter

    International Nuclear Information System (INIS)

    Hankins, D.E.; Homann, S.; Westermark, J.

    1987-01-01

    We developed a personnel neutron dosimetry system based on the electrochemical etching of CR-39 plastic at elevated temperatures. The doses obtained using this dosimeter system are more accurate than those obtained using other dosimetry systems, especially when varied neutron spectra are encountered. This CR-39 dosimetry system does not have the severe energy dependence that exists with albedo neutron dosimeters or the fading and reading problems encountered with NTA film. 3 refs., 4 figs

  18. Perfection of the individual photographic emulsion dosimeter

    International Nuclear Information System (INIS)

    Soudain, G.

    1960-01-01

    A photographic dosimeter making possible the measurement of γ radiation doses of from 10 mr up to 800 r by means of 3 emulsion bands of varying sensitivity stuck to the same support is described. The dosimeter has also a zone for marking and a test film insensitive to radiation. This requires a photometric measurement by diffuse reflection an d makes it possible to measure doses with an accuracy of 20 per cent. (author) [fr

  19. Dosimetric characteristics of PASSAG as a new polymer gel dosimeter with negligible toxicity

    Science.gov (United States)

    Farhood, Bagher; Abtahi, Seyed Mohammad Mahdi; Geraily, Ghazale; Ghorbani, Mehdi; Mahdavi, Seied Rabi; Zahmatkesh, Mohammad Hasan

    2018-06-01

    Despite many advantages of polymer gel dosimeters, their clinical use is only not realized now. Toxicity of polymer gel dosimeters can be considered as one of their main limitations for use in routine clinical applications. In the current study, a new polymer gel dosimeter is introduced with negligible toxicity. For this purpose, 2-Acrylamido-2-Methy-1-PropaneSulfonic acid (AMPS) sodium salt monomer was replaced instead of acrylamide monomer used in PAGAT gel dosimeter by using %6 T and %50 C to the gel formula and the new formulation is called PASSAG (Poly AMPS Sodium Salt and Gelatin) polymer gel dosimeter. The irradiation of gel dosimeters was carried out using a Co-60 therapy machine. MRI technique was used to quantify the dose responses of the PASSAG gel dosimeter. Then, the MRI responses (R2) of the gel dosimeter was analyzed at different dose values, post-irradiation times, and scanning temperatures. The results showed that the new gel formulation has a negligible toxicity and it is also eco-friendly. In addition, carcinogenicity and genetic toxicity tests are negative for the monomer used in PASSAG. The radiological properties of PASSAG gel dosimeter showed that this substance can be considered as a soft tissue/water equivalent material. Furthermore, dosimetric evaluation of the new polymer gel dosimeter revealed an excellent linear R2-dose response in the evaluated dose range (0-15 Gy). The R2-dose sensitivity and dose resolution of PASSAG gel dosimeter were 0.081 s-1Gy-1 (in 0-15 Gy dose range) and 1 Gy (in 0-10 Gy dose range), respectively. Moreover, it was shown that the R2-dose sensitivity and dose resolution of the new gel dosimeter improves over time after irradiation. It was also found that the R2 response of the PASSAG gel dosimeter has less dependency to the 18, 20, and 24 °C scanning temperature in comparison to that of room temperature (22 °C).

  20. Antioxidant effect of green tea on polymer gel dosimeter

    International Nuclear Information System (INIS)

    Samuel, E J J; Sathiyaraj, P; Deena, T; Kumar, D S

    2015-01-01

    Extract from Green Tea (GTE) acts as an antioxidant in acrylamide based polymer gel dosimeter. In this work, PAGAT gel was used for investigation of antioxidant effect of GTE.PAGAT was called PAGTEG (Polyacrylamide green tea extract gel dosimeter) after adding GTE. Free radicals in water cause pre polymerization of polymer gel before irradiation. Polyphenols from GTE are highly effective to absorb the free radicals in water. THPC is used as an antioxidant in polymer gel dosimeter but here we were replaced it by GTE and investigated its effect by spectrophotometer. GTE added PAGAT samples response was lower compared to THPC added sample. To increase the sensitivity of the PAGTEG, sugar was added. This study confirmed that THPC was a good antioxidant for polymer gel dosimeter. However, GTE also can be used as an antioxidant in polymer gel if use less quantity (GTE) and add sugar as sensitivity enhancer

  1. Optimizing the sensitivity and radiological properties of the PRESAGE® dosimeter using metal compounds

    International Nuclear Information System (INIS)

    Alqathami, Mamdooh; Blencowe, Anton; Qiao, Greg; Adamovics, John; Geso, Moshi

    2012-01-01

    The aim of this study is to investigate the radiation-modifying effects of incorporating commercially available bismuth-, tin- and zinc-based compounds in the composition of the PRESAGE ® dosimeter, and the feasibility of employing such compounds for radiation dose enhancement. Furthermore, we demonstrate that metal compounds can be included in the formulation to yield water-equivalent PRESAGE ® dosimeters with enhanced dose response. Various concentrations of the metal compounds were added to a newly developed PRESAGE ® formulation and the resulting dosimeters were irradiated with 100 kV and 6 MV photon beams. A comparison between sensitivity and radiological properties of the PRESAGE ® dosimeters with and without the addition of metal compounds was carried out. Optical density changes of the dosimeters before and after irradiation were measured using a spectrophotometer. In general, when metal compounds were incorporated in the composition of the PRESAGE ® dosimeter, the sensitivity of the dosimeters to radiation dose increased depending on the type and concentration of the metal compound, with the bismuth compound showing the highest dose enhancement factor. In addition, these metal compounds were also shown to improve the retention of the post-response absorption value of the PRESAGE ® dosimeter over a period of 2 weeks. Thus, incorporating 1–3 mM (ca. 0.2 wt%) of any of the three investigated metal compounds in the composition of the PRESAGE ® dosimeter is found to be an efficient way to enhance the sensitivity of the dosimeter to radiation dose and stabilize its post-response for longer times. Furthermore, the addition of small amounts of the metal compounds also accelerates the polymerization of the PRESAGE ® dosimeter precursors, significantly reducing the fabrication time. Finally, a novel water-equivalent PRESAGE ® dosimeter formula optimized with metal compounds is proposed for clinical use in both kilovoltage and megavoltage radiotherapy

  2. Transition in occupational radiation exposure monitoring methods in diagnostic and interventional radiology

    International Nuclear Information System (INIS)

    Loennroth, N.; Hirvonen-Kari, M.; Timonen, M.; Savolainen, S.; Kortesniemi, M.

    2008-01-01

    Radiation exposure monitoring is a traditional keystone of occupational radiation safety measures in medical imaging. The aim of this study was to review the data on occupational exposures in a large central university hospital radiology organisation and propose changes in the radiation worker categories and methods of exposure monitoring. An additional objective was to evaluate the development of electronic personal dosimeters and their potential in the digitised radiology environment. The personal equivalent dose of 267 radiation workers (116 radiologists and 151 radiographers) was monitored using personal dosimeters during the years 2006-2010. Accumulated exposure monitoring results exceeding the registration threshold were observed in the personal dosimeters of 73 workers (59 radiologists' doses ranged from 0.1 to 45.1 mSv; 14 radiographers' doses ranged from 0.1 to 1.3 mSv). The accumulated personal equivalent doses are generally very small, only a few angiography radiologists have doses >10 mSv per 5 y. The typical effective doses are -1 and the highest value was 0.3 mSv (single interventional radiologist). A revised categorisation of radiation workers based on the working profile of the radiologist and observed accumulated doses is justified. Occupational monitoring can be implemented mostly with group dosimeters. An active real-time dosimetry system is warranted to support radiation protection strategy where optimisation aspects, including improving working methods, are essential. (authors)

  3. Studies on α-Al2O3: C based optically stimulated luminescence badge for eye lens monitoring applications

    International Nuclear Information System (INIS)

    Kumar, Munish; Kulkarni, M.S.; Ratna, P.; Gaikwad, N.; Tripathi, S.M.; Sharma, S.D.; Babu, D.A.R.; Bhatnagar, Amit; Muthe, K.P.; Sharma, D.N.

    2014-01-01

    The prototype two element eye-lens dosimeter badge based on indigenously developed α-Al 2 O 3 : C optically stimulated luminescence dosimeter was investigated comprehensively for its suitability for eye-lens monitoring applications. The badge is calibrated to measure the eye-lens dose in terms of H p (3). The minimum measurable dose using the eye-lens dosimeter badge is observed to be ∼ 35 μSv. This prototype eye-lens dosimeter badge was found to be suitable for measuring doses from X-rays, beta and gamma radiations to the eye-lens. The satisfactory performance of the prototype two element eye-lens dosimeter badge along with its attractive features such as multiple readout, less processing time, very good beta response uniquely position it for monitoring the eye-lens dose are presented. (author)

  4. Development of new chemical dosimeter for low dose range

    International Nuclear Information System (INIS)

    Mhatre, Sachin G.V.; Adhikari, S.

    2012-01-01

    Accurate measurement of low dose radiation in complex systems is of utmost importance in radiation biology and related areas. Ferrous Benzoic acid Xylenol orange (FBX) system is being widely used for measurement of low dose gamma radiation because of its reproducibility and precision. However, an additional step, i.e., dissolution of benzoic acid in water at higher temperature followed by cooling at room temperature is involved for the preparation of this dosimeter. This makes it inconvenient as a ready to use dosimeter. In the present work, the organic molecule, sorbitol has been used for measurement of low doses of radiation. The advantages of using sorbitol are its ready availability and instantaneous water solubility. Owing to its dissolution at room temperature, possible errors those are involved in calculation of dose due to thermal oxidation of ferrous ions during preparation of the FBX dosimetric solution could be made insignificant in the proposed dosimeter. In the present system, sorbitol acts as radiolytic sensitizer for the oxidation of ferrous ion, and xylenol orange forms a 1:1 complex specifically with ferric ions. Thus, the analytical detection limit of ferric ions is enhanced compared to other systems. Final composition of the dosimetric solution is; 0.5 mol/m 3 xylenol orange, 10 mol/m 3 sorbitol and 0.2 mol/m 3 ferrous ion in 50 mol/m 3 sulfuric acid. Radiolytic sensitization in combination with analytical enhancement of the ferrous based system, allows us to measure radiation dose in the range of 0.05 Gy–12 Gy with ease and high reproducibility.

  5. An NMR relaxometry and gravimetric study of gelatin-free aqueous polyacrylamide dosimeters

    International Nuclear Information System (INIS)

    Babic, Steven; Schreiner, L John

    2006-01-01

    In conformal radiation therapy, a high dose of radiation is given to a target volume to increase the probability of cure, and care is taken to minimize the dose to surrounding healthy tissue. The techniques used to achieve this are very complicated and the precise verification of the resulting three-dimensional (3D) dose distribution is required. Polyacrylamide gelatin (PAG) dosimeters with magnetic resonance imaging and optical computed tomography scanning provide the required 3D dosimetry with high spatial resolution. Many basic studies have characterized these chemical dosimeters that polymerize under irradiation. However, the investigation of the fundamental properties of the radiation-induced polymerization in PAG dosimeters is complicated by the presence of the background gelatin matrix. In this work, a gelatin-free model system for the study of the basic radiation-induced polymerization in PAG dosimeters has been developed. Experiments were performed on gelatin-free dosimeters, named aqueous polyacrylamide (APA) dosimeters, containing equal amounts of acrylamide and N,N'-methylene-bisacrylamide. The APA dosimeters were prepared with four different total monomer concentrations (2, 4, 6 and 8% by weight). Nuclear magnetic resonance (NMR) spin-spin and spin-lattice proton relaxation measurements at 20 MHz, and gravimetric analyses performed on all four dosimeters, show a continuous degree of polymerization over the dose range of 0-25 Gy. The developed NMR model explains the relationship observed between the relaxation data and the amount of crosslinked polymer formed at each dose. This model can be extended with gelatin relaxation data to provide a fundamental understanding of radiation-induced polymerization in the conventional PAG dosimeters

  6. Development and characterization of real-time wide-energy range personal neutron dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Nakamura, Takashi; Tsujimura, Norio (Tohoku Univ., Sendai (Japan). Cyclotron and Radioisotope Center); Yamano, Toshiya; Suzuki, Toshikazu; Okamoto, Eisuke

    1994-04-01

    The authors developed a real-time personal neutron dosimeter which could give neutron dose equivalent over wide energy region from thermal to 10 odd MeV by using 2 silicon detectors, fast neutron sensor and slow neutron sensor. The energy response of this dosimeter was evaluated under thermal neutron field, monoenergetic neutron field between 200 keV and 15 MeV, and moderated [sup 252]Cf neutron field. The neutron dose equivalent was estimated by adding neutron dose equivalent below 1 MeV given by slow neutron sensor and that above 1 MeV by fast neutron sensor. It was verified from various field tests that this dosimeter is able to give neutron dose equivalent within a factor of 2 margin of accuracy in reactor, accelerator, fusion research and nuclear fuel handling facilities. This dosimeter has more than one order higher sensitivity than conventional personal neutron dosimeters and is insensitive to [gamma]-rays up to about 500 mSv/h. This dosimeter will soon be commercially available as a personal dosimeter which gives neutron and [gamma]-ray dose equivalents simultaneously by installing [gamma]-ray silicon sensor. (author).

  7. Dental tissue as a thermoluminescence dosimetry dosimeter

    International Nuclear Information System (INIS)

    Solaimani, F.; Zahmatkesh, M.H.; Akhlaghpoor, Sh.

    2003-01-01

    Background: Thermoluminescence dosimetry is one of the dosimetry procedures used widely as routine and personal dosimeters. In order to extend this kind of dosimeters, dental tissue has been examined and was found promising as a Thermoluminescence Dosimetry dosimeter. Materials and Methods: In this study, 70 health teeth were collected. The only criterion, wich was considered for selection of the teeth, was the healthiness of them regardless of age and gender of the donors. All collected samples were washed and cleaned and milled uniformly. The final powder had a uniform grain size between 100-300 micrometer. The sample was divided into four groups. Group A and B were used for measurement of density and investigation of variation of thermoluminescent characteristics with temperature respectively. Groups C and D were used for investigation of variation of thermoluminescent intensity with dose and fading of this intensity with time. In all cases the results obtained with dental tissue were compared to a standard LiF, thermoluminescence dosimetry dosimeter. Results: It was found that, average density of the dental tissue was 1.570 g/cm 3 , which is comparable to density of LiF, which is 1.612g/cm 3 . It was also concluded that the range of 0-300 d ig C , dental tissue has a simple curve with two specific peaks at 140 and 25 d ig C respectively. The experiment also showed that, the variation of relative intensity versus dose is linear in the range of 0.04-0.1 Gy. The fading rate of dental tissue is higher than LiF but still in the acceptable range (14% per month in compare to 5.2% per month). Conclusion: Dental tissue as a natural dosimeter is comparable with Thermoluminescence Dosimetry and can be used in accidental events with a good approximation

  8. Experimental evaluation of a MOSFET dosimeter for proton dose measurements

    International Nuclear Information System (INIS)

    Kohno, Ryosuke; Nishio, Teiji; Miyagishi, Tomoko; Hirano, Eriko; Hotta, Kenji; Kawashima, Mitsuhiko; Ogino, Takashi

    2006-01-01

    The metal oxide semiconductor field-effect transistor (MOSFET) dosimeter has been widely studied for use as a dosimeter for patient dose verification. The major advantage of this detector is its size, which acts as a point dosimeter, and also its ease of use. The commercially available TN502RD MOSFET dosimeter manufactured by Thomson and Nielsen has never been used for proton dosimetry. Therefore we used the MOSFET dosimeter for the first time in proton dose measurements. In this study, the MOSFET dosimeter was irradiated with 190 MeV therapeutic proton beams. We experimentally evaluated dose reproducibility, linearity, fading effect, beam intensity dependence and angular dependence for the proton beam. Furthermore, the Bragg curve and spread-out Bragg peak were also measured and the linear-energy transfer (LET) dependence of the MOSFET response was investigated. Many characteristics of the MOSFET response for proton beams were the same as those for photon beams reported in previous papers. However, the angular MOSFET responses at 45, 90, 135, 225, 270 and 315 degrees for proton beams were over-responses of about 15%, and moreover the MOSFET response depended strongly on the LET of the proton beam. This study showed that the angular dependence and LET dependence of the MOSFET response must be considered very carefully for quantitative proton dose evaluations

  9. Early development and characterization of a DNA-based radiation dosimeter

    Science.gov (United States)

    Avarmaa, Kirsten A.

    It is the priority of first responders to minimize damage to persons and infrastructure in the case of a nuclear emergency due to an accident or deliberate terrorist attack -- if this emergency includes a radioactive hazard, first responders require a simple-to-use, accurate and complete dosimeter for radiation protection purposes in order to minimize the health risk to these individuals and the general population at large. This work consists of the early evaluation of the design and performance of a biologically relevant dosimeter which uses DNA material that can respond to the radiation of any particle type. The construct consists of fluorescently tagged strands of DNA. The signalling components of this dosimeter are also investigated for their sensitivity to radiation damage and light exposure. The dual-labelled dosimeter that is evaluated in this work gave a measurable response to gamma radiation at dose levels of 10 Gy for the given detector design and experimental setup. Further testing outside of this work confirmed this finding and indicated a working range of 100 mGy to 10 Gy using a custom-built fluorimeter as part of a larger CRTI initiative. Characterization of the chromatic components of the dosimeter showed that photobleaching is not expected to have an effect on dosimeter performance, but that radiation can damage the non-DNA signalling components at higher dose levels, although this damage is minimal at lower doses over the expected operating ranges. This work therefore describes the early steps in the quantification of the behaviour of the DNA dosimeter as a potential biologically-based device to measure radiation dose.

  10. Dosimetry of blood irradiation using an alanine/ESR dosimeter

    International Nuclear Information System (INIS)

    Chen, F.; Covas, D.T.; Baffa, O.

    2001-01-01

    A batch of 80 DL-alanine dosimeters was supplied to Hemocentro of the Hospital and Clinics of Faculdade de Medicina de Ribeirao Preto (HC-FMRP) SP, Brazil for the purpose of quality control of the radiation dose delivered to blood bags. The irradiation was made using two (40x40) cm 2 parallel opposed radiation fields each with 80 cm of source to surface distance in the Radiotherapy Section of HC-FMRP with the 60 Co teletherapy unit. The calculated radiation absorbed dose at the center of the box was 20 Gy. The dosimeter readings were performed using a Varian E-4 ESR Spectrometer operating in X-band. For the 80 dosimeters and over the irradiation volume throughout a blood bag, the minimum and maximum doses were 14 and 23 Gy, respectively. The mean dose was (18±2) Gy (1σ), and the coefficient of variability was 11.1%. Alanine dosimeters demonstrated easy handling, good precision and adequate sensitivity for this application

  11. Electret dosimeter utilizing gas multiplication

    International Nuclear Information System (INIS)

    Ikeya, M.; Miki, T.

    1980-01-01

    It was found that the high electric field around the surface of an electret leads to cascade multiplication of the ionization process in a surrounding gas. Very sensitive charge decay constants of the order of 1mrad, were obtained for electrets composed of polyvinyliden fluoride or teflon polymers. The reduced charge is stable and can be utilized in personnel dosimetry. A simple pocket chamber dosimeter is described consisting of a small speaker or buzzer, a cylindrical chamber filled with air, argon or other gases, a polymer thermoelectret foil and an electrode. The sonic vibration of the foil induces an alternating charge on the electrode which is amplified and detected. The feasibility of this dosimeter and its shock and vibration resistance have been demonstrated. (author)

  12. Monte Carlo simulation experiments on box-type radon dosimeter

    International Nuclear Information System (INIS)

    Jamil, Khalid; Kamran, Muhammad; Illahi, Ahsan; Manzoor, Shahid

    2014-01-01

    Epidemiological studies show that inhalation of radon gas ( 222 Rn) may be carcinogenic especially to mine workers, people living in closed indoor energy conserved environments and underground dwellers. It is, therefore, of paramount importance to measure the 222 Rn concentrations (Bq/m 3 ) in indoors environments. For this purpose, box-type passive radon dosimeters employing ion track detector like CR-39 are widely used. Fraction of the number of radon alphas emitted in the volume of the box type dosimeter resulting in latent track formation on CR-39 is the latent track registration efficiency. Latent track registration efficiency is ultimately required to evaluate the radon concentration which consequently determines the effective dose and the radiological hazards. In this research, Monte Carlo simulation experiments were carried out to study the alpha latent track registration efficiency for box type radon dosimeter as a function of dosimeter’s dimensions and range of alpha particles in air. Two different self developed Monte Carlo simulation techniques were employed namely: (a) Surface ratio (SURA) method and (b) Ray hitting (RAHI) method. Monte Carlo simulation experiments revealed that there are two types of efficiencies i.e. intrinsic efficiency (η int ) and alpha hit efficiency (η hit ). The η int depends upon only on the dimensions of the dosimeter and η hit depends both upon dimensions of the dosimeter and range of the alpha particles. The total latent track registration efficiency is the product of both intrinsic and hit efficiencies. It has been concluded that if diagonal length of box type dosimeter is kept smaller than the range of alpha particle then hit efficiency is achieved as 100%. Nevertheless the intrinsic efficiency keeps playing its role. The Monte Carlo simulation experimental results have been found helpful to understand the intricate track registration mechanisms in the box type dosimeter. This paper explains that how radon

  13. Monte Carlo simulation experiments on box-type radon dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Jamil, Khalid, E-mail: kjamil@comsats.edu.pk; Kamran, Muhammad; Illahi, Ahsan; Manzoor, Shahid

    2014-11-11

    Epidemiological studies show that inhalation of radon gas ({sup 222}Rn) may be carcinogenic especially to mine workers, people living in closed indoor energy conserved environments and underground dwellers. It is, therefore, of paramount importance to measure the {sup 222}Rn concentrations (Bq/m{sup 3}) in indoors environments. For this purpose, box-type passive radon dosimeters employing ion track detector like CR-39 are widely used. Fraction of the number of radon alphas emitted in the volume of the box type dosimeter resulting in latent track formation on CR-39 is the latent track registration efficiency. Latent track registration efficiency is ultimately required to evaluate the radon concentration which consequently determines the effective dose and the radiological hazards. In this research, Monte Carlo simulation experiments were carried out to study the alpha latent track registration efficiency for box type radon dosimeter as a function of dosimeter’s dimensions and range of alpha particles in air. Two different self developed Monte Carlo simulation techniques were employed namely: (a) Surface ratio (SURA) method and (b) Ray hitting (RAHI) method. Monte Carlo simulation experiments revealed that there are two types of efficiencies i.e. intrinsic efficiency (η{sub int}) and alpha hit efficiency (η{sub hit}). The η{sub int} depends upon only on the dimensions of the dosimeter and η{sub hit} depends both upon dimensions of the dosimeter and range of the alpha particles. The total latent track registration efficiency is the product of both intrinsic and hit efficiencies. It has been concluded that if diagonal length of box type dosimeter is kept smaller than the range of alpha particle then hit efficiency is achieved as 100%. Nevertheless the intrinsic efficiency keeps playing its role. The Monte Carlo simulation experimental results have been found helpful to understand the intricate track registration mechanisms in the box type dosimeter. This paper

  14. Individual monitoring of medical staff working in interventional radiology in Switzerland using double dosimetry

    International Nuclear Information System (INIS)

    Damet, J.; Bailat, C.; Bize, P.; Buchillier, Th.; Tosic, M.; Verdun, F.R.; Baechler, S.

    2011-01-01

    Physicians who frequently perform fluoroscopic examinations are exposed to high intensity radiation fields. The exposure monitoring is performed with a regular personal dosimeter under the apron in order to estimate the effective dose. However, large parts of the body are not protected by the apron (e.g. arms, head). Therefore, it is recommended to wear a supplemental dosimeter over the apron to obtain a better representative estimate of the effective dose. The over-apron dosimeter can also be used to estimate the eye lens dose. The goal of this study was to investigate the relevance of double dosimetry in interventional radiology. First the calibration procedure of the dosimeters placed over the apron was tested. Then, results of double dosimetry during the last five years were analyzed. We found that the personal dose equivalent measured over a lead apron was underestimated by ∼20% to ∼40% for X-ray beam qualities used in radiology. Measurements made over five-year period confirm that the use of a single under-apron dosimeter is inadequate for personnel monitoring. Relatively high skin dose (>10 mSv/month) would have remained undetected without a second dosimeter placed on the apron.

  15. Telepositional portable real time radiation monitoring system

    International Nuclear Information System (INIS)

    Talpalariu, Jeni; Matei, Corina; Popescu, Oana

    2010-01-01

    Technology development for complex portable networks is on going to meet the area dosimetry challenge, improving the basic design using new telepositional GPS satellite methods and GSM terrestrial civil radio transmission networks. The system and devices proposed overcome the limitations of fixed and portable dosimeters, providing wireless real time radiations data and geospatial information's means, using many portable dosimeter stations and a mobile dosimeter computerised central console. (authors)

  16. Evaluation of aluminum oxide dosimeters using OSL technique in dosimetry of clinical photon beams on volumetric modulated arc treatment

    International Nuclear Information System (INIS)

    Villani, Daniel

    2017-01-01

    Treatment using Volumetric Modulated Arc Radiation Therapy is the most modern modality of conformational radiotherapy so that, with the overlapping of several fields, the dose distributions provide a perfect conformation to the tumor, reducing the probability of complications in adjacent normal tissues. In this sense, many efforts are being invested to improve dose distribution compliance as well as the integration of imaging techniques for tumor screening and correction of inter and intrafraction variations. To this end, an intensive monitoring of the quality of the processes and a quality assurance program are fundamental for patient safety and compliance with current legislation; besides the use of different dosimetry methodologies for intercomparison and validation of the results. The aim of this study is to evaluate and compare the performance of aluminum oxide (Al_2O_3:C) OSL dosimeters manufactured by Landauer Inc. with those produced by Rexon™ in the dosimetry of high energy photon clinical bundles used in Volumetric Modulated Arc Therapy - (VMAT) using different simulating objects. The dosimeters were characterized for gamma radiation of the "6"0Co and for clinical photon beams of 6 MV typical of treatments by VMAT under conditions of electronic equilibrium and maximum dose respectively. Performance tests of the TL and OSL readers used and repeatability of the samples were evaluated. After all tests, the dosimeters were irradiated in the simulation of different radiotherapy treatments by VMAT and their responses compared to the planning system. All types of dosimeters presented satisfactory results in verifying the doses of this type of planning simulation. The Al_2O_3:C dosimeters presented compatible results and validated by the other dosimeters and ionization chamber. Regarding the best technique, the OSL InLight commercial system presents greater practicality and versatility for use and application in the clinical routine. (author)

  17. 10 CFR 36.55 - Personnel monitoring.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 1 2010-01-01 2010-01-01 false Personnel monitoring. 36.55 Section 36.55 Energy NUCLEAR REGULATORY COMMISSION LICENSES AND RADIATION SAFETY REQUIREMENTS FOR IRRADIATORS Operation of Irradiators § 36.55 Personnel monitoring. (a) Irradiator operators shall wear a personnel dosimeter that is...

  18. Angular dependence of the nanoDot OSL dosimeter

    OpenAIRE

    Kerns, James R.; Kry, Stephen F.; Sahoo, Narayan; Followill, David S.; Ibbott, Geoffrey S.

    2011-01-01

    Purpose: Optically stimulated luminescent detectors (OSLDs) are quickly gaining popularity as passive dosimeters, with applications in medicine for linac output calibration verification, brachytherapy source verification, treatment plan quality assurance, and clinical dose measurements. With such wide applications, these dosimeters must be characterized for numerous factors affecting their response. The most abundant commercial OSLD is the InLight∕OSL system from Landauer, Inc. The purpose of...

  19. Determination of radiation direction in environmental monitoring

    International Nuclear Information System (INIS)

    Campos, Vicente de Paulo de; Moura, Eduardo S.; Rocha, Felicia D.G.; Manzoli, Jose Eduardo

    2009-01-01

    The assessment of environmental exposure has been performed in Brazil using the thermoluminescence technique at Thermoluminescence Dosimetry Laboratory (LDT), at Nuclear and Energetic Research Institute (IPEN/CNEN-SP). To carry out these measurements, several thermoluminescent dosimeters (TLD's) were used to measure the expose. In this procedure, very few information of direction where the radiation came from is available. A vague supposition about the direction from where the radiation came from could be inferred only by evaluation of multiple dosimeters displaced at entire region of monitoring, but this demand to much effort or sometimes become impractical for certain situations. In this work, a single device is used to provide information about the direction from where the radiation came through. This device is called directional dosimeter (DD). Using more than one DD it is possible to reduce the uncertainty of the measurements and determine the radiation source position. The DD basically consists of a regular solid with high effective atomic number, where one TLD is positioned at each face. The DD allows evaluating the environmental exposure and the direction of the radiation by a simple vector sum. At each face of the DD, it is associated an orthogonal vector, and modulus of this vector represents the correspond exposure measured by the TLD. The direction of the radiation source is the sum of these faces vectors. The prototype used in this work was a lead cube with six TLDs of CaSO 4 :Dy/Teflon. The TLDs have high sensibility and are already used in area, environmental and personal monitoring. The measurements had shown the correct environmental exposure and a good indication of the radiation direction. (author)

  20. An automatic evaluation system for NTA film neutron dosimeters

    CERN Document Server

    Müller, R

    1999-01-01

    At CERN, neutron personal monitoring for over 4000 collaborators is performed with Kodak NTA films, which have been shown to be the most suitable neutron dosimeter in the radiation environment around high-energy accelerators. To overcome the lengthy and strenuous manual scanning process with an optical microscope, an automatic analysis system has been developed. We report on the successful automatic scanning of NTA films irradiated with sup 2 sup 3 sup 8 Pu-Be source neutrons, which results in densely ionised recoil tracks, as well as on the extension of the method to higher energy neutrons causing sparse and fragmentary tracks. The application of the method in routine personal monitoring is discussed. $9 overcome the lengthy and strenuous manual scanning process with an optical microscope, an automatic analysis system has been developed. We report on the successful automatic scanning of NTA films irradiated with /sup 238/Pu-Be source $9 discussed. (10 refs).

  1. Anthracene dosimeter characterization under radiotherapy photons

    International Nuclear Information System (INIS)

    Czelusniak, Caroline

    2011-01-01

    New radiotherapy techniques such as intensity-modulated radiation therapy and stereotactic radiosurgery have increased the need for dosimeters that can provide measurements in real time with high spatial resolution. Organic scintillation dosimeters are able to measure with accuracy small radiation fields and fields with high gradients, besides having advantages such as water and soft tissue equivalence and the possibility to be used in vivo. Anthracene is an organic scintillator crystal with the highest known scintillation efficiency among organic scintillation materials. The objective of this work is to characterize the anthracene as a dosimeter under radiotherapy photons energies, analysing its signal against average granulosity, intern capsule diameter, absorbed dose, absorbed dose rate, photon energy and its spatial resolution; with the last one analysed under three methods (edge spread function, line spread function and modulation transfer function). The photons energies used were 1.25 MeV ( 60 Co), 0.661 MeV ( 137 Cs) and X-rays (effective energies of 28.4; 46.5; 48.5; 94.0 e 106.0 keV). The scintillation detection system consisted of an optical fiber with one end attached to the anthracene capsule and the other to a photomultiplier tube maintained by power supply followed by an electrometer. Once Cerenkov radiation occurs in the optical fiber, it was removed from the total scintillation signal trough the subtraction of the signal, taken irradiating the optical fiber without the anthracene attached to one of its extremity. From results obtained, one can infer that the dosimeter signal increases proportionally with average granulosity and intern capsule diameter. The signal is linearly dependent of absorbed dose, linearly dependent of low photons energies and independent for high photons energies, as well as independent of the absorbed dose rate. From the spatial resolution values obtained it was possible to infer that the one obtained through modulation

  2. Automatic dosimeter for kerma measurement based on commercial PIN photo diodes

    International Nuclear Information System (INIS)

    Kushpil, V.; Kushpil, S.; Huna, Z.

    2011-01-01

    A new automatic dosimeter for measurement of radiation dose from neutron and ionization radiation is presented. The dosimeter (kerma meter) uses commercial PIN diodes with long base as its active element. Later it provides a maximal dependence of the minority carriers life time versus absorbed dose. The characteristics of the dosimeter were measured for several types of commercial diodes. Device can be useful in many environmental or industrial applications. (authors)

  3. Use of normoxic polymer gel dosimeters for measuring diagnostic doses on CT scanners

    International Nuclear Information System (INIS)

    Hill, B; Venning, A J; Baldock, C

    2004-01-01

    X-ray CT has been used to evaluate polymer gel dosimeters for dose response in the therapeutic dose range. This method of polymer gel dosimeter evaluation has been shown to be useful for instance in the comparison of complex sterotactic field distributions with treatment plans. Image averaging and subtraction techniques are used for noise reduction in polymer gel dosimeters resulting in the delivery of several CT slices across the polymer gel dosimeters. It was a logical progression to evaluate normoxic polymer gel dosimeters with optimized CT scanning protocols. During these investigations it was found that unirradiated regions in irradiated normoxic polymer gel dosimetry phantoms polymerised possibly as a result of the evaluation using CT. This prompted an investigation of the CT diagnostic dose response of the normoxic polymer gel dosimeter in order to determine the dose contribution when evaluated using a CT scanner. Having established that there was an effect on the normoxic polymer gel dosimeter when evaluating with a CT scanner the suitability of these gels in the determination of CT diagnostic dose measurement was further investigated

  4. Development of a miniaturized watch-type dosimeter using a silicon printed-circuit board

    International Nuclear Information System (INIS)

    Ishikura, Takeshi; Sakamaki, Tsuyoshi; Matsumoto, Iwao; Aoyama, Kei; Nakamura, Takashi

    2008-01-01

    The electrical personal dosimeter using a silicon semiconductor sensor has the advantage of real time response and alarm function, which can prevent unexpected over-exposure. We tried to develop a miniaturized watch-type dosimeter by incorporating the silicon semiconductor sensor on a silicon printed-circuit board. Thin film resistors, capacitors and wiring patterns are formed on a downsized printed-circuit board. Electronic parts including transistors are mounted by soldering on the silicon printed-circuit board. The dosimeter is further miniaturized by downsizing the amplifier circuit, the semiconductor radiation sensor, the power supply circuit, setting parts and alarm part. The performance of the developed dosimeter was evaluated with respect to the gamma-ray spectra, angular dependence and linearity to dose equivalent rate, and it was confirmed that this dosimeter has the performance equivalent to a commercially available electrical personal dosimeter. (author)

  5. Comprehensive Angular Response Study of LLNL Panasonic Dosimeter Configurations and Artificial Intelligence Algorithm

    Energy Technology Data Exchange (ETDEWEB)

    Stone, D. K. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2017-06-30

    In April of 2016, the Lawrence Livermore National Laboratory External Dosimetry Program underwent a Department of Energy Laboratory Accreditation Program (DOELAP) on-site assessment. The assessment reported a concern that the study performed in 2013 Angular Dependence Study Panasonic UD-802 and UD-810 Dosimeters LLNL Artificial Intelligence Algorithm was incomplete. Only the responses at ±60° and 0° were evaluated and independent data from dosimeters was not used to evaluate the algorithm. Additionally, other configurations of LLNL dosimeters were not considered in this study. This includes nuclear accident dosimeters (NAD) which are placed in the wells surrounding the TLD in the dosimeter holder.

  6. Dosimetric Characteristics of Radio-Photoluminescent Glass Dosimeters for Medical Applications: Linearity

    Energy Technology Data Exchange (ETDEWEB)

    Shehzadi, N. N.; Jeong, J. P.; Kim, B. C.; Kim, I. J.; Yi, C. Y. [Center for Ionizing Radiation, Korea Research Institute of Standards and Science, Seoul (Korea, Republic of)

    2017-04-15

    Radio-photoluminescent glass dosimeter (GD) has advantage of non-destructive reading process, negligible fading and superior radiation detection characteristics than other personal dosimeters like thermoluminescent dosimeters (TLD) or film dosimeters. In this study, one dosimetric characteristic of GDs, dose linearity was evaluated with two different approaches: one for each set of GDs selected and another for a batch of them using accumulated doses of same set of GDs and GDs in a batch, respectively. Within a dose range upto 10 Gy, not only each set of GDs but also a batch of them showed excellent linearity. Within a dose range upto 10 Gy, not only each set of GDs but also a batch of them showed excellent linearity.

  7. Electronic Personal Dosimeters Open a New Dimension in Radon Dose Management

    International Nuclear Information System (INIS)

    Streil, T.; Oeser, V.

    2001-01-01

    Full text: Electronic Radon dosimeters enable the installation of completely automated dose management systems in Radon exposed areas for the first time. In opposition to passive dosimeters, the actual dose value will be displayed online. The alarm function indicates the reaching of the permissible doses and prevents exceeding of given levels. The immediate availability of all user- and measurement information leads to a new level of quality assurance within complex dose management systems. Furthermore, the sequentially stored data give an exact assignment of concentration and dose values to the real time and location. This information is very important for staff planning and the assessment of Radon affected objects (local dosimeters). The measurement of Radon concentration is based on the alpha spectroscopy. The gas diffuses through a membrane into the measurement chamber. Progeny inside the chamber ionised after decay will be collected at the detector surface forced by the electric field. All incoming events will be processed by a Multi Channel Analyser (MCA). A integral spectrum and a record of five peak-areas (each assigned to a single nuclide) at every time step will be stored for computing concentration and dose values. The sensitivity of the device was determined to 0.25 counts/(minkBq/m 3 ). An average concentration of 200 Bq/m 3 during an eight-hours work day gives an error of ±20%. The response time (95% of final value) only depends on half live times of 218 Po and 214 Po (10 minutes using fast mode, 2 hours in slow mode) and is not affected by the diffusion membrane. Further tests at high levels (up to several MBq/m 3 ) were carried out successfully during soil gas and water measurements. (author)

  8. Radiation-induced coloration of nitro blue tetrazolium gel dosimeter for low dose applications

    International Nuclear Information System (INIS)

    Abdel-Fattah, A.A.; Beshir, W.B.; Hassan, H.M.; Soliman, Y.S.

    2017-01-01

    A radiochromic sensor of nitro blue tetrazolium (NBT) in gelatin was evaluated as a new gel dosimeter for radiation applications. The NBT gel has the advantage of visual color change from faint yellow to violet at low absorbed doses (10–1000 Gy). This color change appears as a result of the reduction of NBT to colored formazan then to diformazan species with further increase of absorbed doses. Responses of the gel at different NBT concentrations were analyzed at the absorption maximum centered at 527 nm. An increase of NBT concentrations in the gel enhances the radiation dose sensitivity. Energy dependent study implies the tissue equivalency of the gel in the energy range of 0.15–20 MeV. Dependence of the gel response on irradiation temperature, and color stability before and after irradiation were also studied. The combined uncertainty associated with the dose monitoring (10–1000 Gy) is 6.26% (2σ). Thus, the NBT gel shows its suitability for food irradiation, insect population control, and some food irradiation applications. - Highlights: • Preparation of nitro blue tetrazolium (NBT) gel for the dose range of 10–1000 Gy. • The sensitivity of it increases with increasing NBT concentrations. • The response of irradiated dosimeter is stable after 5 h from irradiation. • The prepared gel dosimeter is a tissue equivalent. • Its combined uncertainty is equal to 6.26% for 10–1000 Gy dose level.

  9. Response of TLD-albedo and nuclear track dosimeters exposed to plutonium sources

    International Nuclear Information System (INIS)

    Brackenbush, L.W.; Baumgartner, W.V.; Fix, J.J.

    1991-12-01

    Neutron dosimetry has been extensively studied at Hanford since the mid-1940s. At the present time, Hanford contractors use thermoluminescent dosimeter (TLD)-albedo dosimeters to record the neutron dose equivalent received by workers. The energy dependence of the TLD-albedo dosimeter has been recognized and documented since introduced at Hanford in 1964 and numerous studies have helped assure the accuracy of dosimeters. With the recent change in Hanford's mission, there has been a significant decrease in the handling of plutonium tetrafluoride, and an increase in the handling of plutonium metal and plutonium oxide sources. This study was initiated to document the performance of the current Hanford TLD-albedo dosimeter under the low scatter conditions of the calibration laboratory and under the high scatter conditions in the work place under carefully controlled conditions at the Plutonium Finishing Plant (PFP). The neutron fields at the PFP facility were measured using a variety of instruments, including a multisphere spectrometer, tissue equivalent proportional counters, and specially calibrated rem meters. Various algorithms were used to evaluate the TLD-albedo dosimeters, and the results are given in this report. Using current algorithms, the dose equivalents evaluated for bare sources and sources with less than 2.5 cm (1 in.) of acrylic plastic shielding in high scatter conditions typical of glove box operations are reasonably accurate. Recently developed CR-39 track etch dosimeters (TEDs) were also exposed in the calibration laboratory and at the PFP. The results indicate that the TED dosimeters are quite accurate for both bare and moderated neutron sources. Until personnel dosimeter is available that incorporates a direct measure of the neutron dose to a person, technical uncertainties in the accuracy of the recorded data will continue

  10. Response of TLD-albedo and nuclear track dosimeters exposed to plutonium sources

    Energy Technology Data Exchange (ETDEWEB)

    Brackenbush, L.W.; Baumgartner, W.V.; Fix, J.J.

    1991-12-01

    Neutron dosimetry has been extensively studied at Hanford since the mid-1940s. At the present time, Hanford contractors use thermoluminescent dosimeter (TLD)-albedo dosimeters to record the neutron dose equivalent received by workers. The energy dependence of the TLD-albedo dosimeter has been recognized and documented since introduced at Hanford in 1964 and numerous studies have helped assure the accuracy of dosimeters. With the recent change in Hanford`s mission, there has been a significant decrease in the handling of plutonium tetrafluoride, and an increase in the handling of plutonium metal and plutonium oxide sources. This study was initiated to document the performance of the current Hanford TLD-albedo dosimeter under the low scatter conditions of the calibration laboratory and under the high scatter conditions in the work place under carefully controlled conditions at the Plutonium Finishing Plant (PFP). The neutron fields at the PFP facility were measured using a variety of instruments, including a multisphere spectrometer, tissue equivalent proportional counters, and specially calibrated rem meters. Various algorithms were used to evaluate the TLD-albedo dosimeters, and the results are given in this report. Using current algorithms, the dose equivalents evaluated for bare sources and sources with less than 2.5 cm (1 in.) of acrylic plastic shielding in high scatter conditions typical of glove box operations are reasonably accurate. Recently developed CR-39 track etch dosimeters (TEDs) were also exposed in the calibration laboratory and at the PFP. The results indicate that the TED dosimeters are quite accurate for both bare and moderated neutron sources. Until personnel dosimeter is available that incorporates a direct measure of the neutron dose to a person, technical uncertainties in the accuracy of the recorded data will continue.

  11. Response of TLD-albedo and nuclear track dosimeters exposed to plutonium sources

    Energy Technology Data Exchange (ETDEWEB)

    Brackenbush, L.W.; Baumgartner, W.V.; Fix, J.J.

    1991-12-01

    Neutron dosimetry has been extensively studied at Hanford since the mid-1940s. At the present time, Hanford contractors use thermoluminescent dosimeter (TLD)-albedo dosimeters to record the neutron dose equivalent received by workers. The energy dependence of the TLD-albedo dosimeter has been recognized and documented since introduced at Hanford in 1964 and numerous studies have helped assure the accuracy of dosimeters. With the recent change in Hanford's mission, there has been a significant decrease in the handling of plutonium tetrafluoride, and an increase in the handling of plutonium metal and plutonium oxide sources. This study was initiated to document the performance of the current Hanford TLD-albedo dosimeter under the low scatter conditions of the calibration laboratory and under the high scatter conditions in the work place under carefully controlled conditions at the Plutonium Finishing Plant (PFP). The neutron fields at the PFP facility were measured using a variety of instruments, including a multisphere spectrometer, tissue equivalent proportional counters, and specially calibrated rem meters. Various algorithms were used to evaluate the TLD-albedo dosimeters, and the results are given in this report. Using current algorithms, the dose equivalents evaluated for bare sources and sources with less than 2.5 cm (1 in.) of acrylic plastic shielding in high scatter conditions typical of glove box operations are reasonably accurate. Recently developed CR-39 track etch dosimeters (TEDs) were also exposed in the calibration laboratory and at the PFP. The results indicate that the TED dosimeters are quite accurate for both bare and moderated neutron sources. Until personnel dosimeter is available that incorporates a direct measure of the neutron dose to a person, technical uncertainties in the accuracy of the recorded data will continue.

  12. High dose potassium-nitrate chemical dosimeter

    International Nuclear Information System (INIS)

    Dorda de Cancio, E.M.; Munoz, S.S.

    1982-01-01

    This dosimeter is used to control 10 kGY-order doses (1 Mrad). Nitrate suffers a radiolitic reduction phenomena, which is related to the given dose. The method to use potassium nitrate as dosimeter is described, as well as effects of the temperature of irradiation, pH, nitrate concentration and post-irradiation stability. Nitrate powder was irradiated at a Semi-Industrial Plant, at Centro Atomico Ezeiza, and also in a Gammacell-220 irradiator. The dose rates used were 2,60 and 1,80 KGY/hour, and the given doses varied between 1,0 and 150 KGY. The uncertainty was +-3% in all the range. (author) [es

  13. Dose determination algorithms for a nearly tissue equivalent multi-element thermoluminescent dosimeter

    International Nuclear Information System (INIS)

    Moscovitch, M.; Chamberlain, J.; Velbeck, K.J.

    1988-01-01

    In a continuing effort to develop dosimetric systems that will enable reliable interpretation of dosimeter readings in terms of the absorbed dose or dose-equivalent, a new multi-element TL dosimeter assembly for Beta and Gamma dose monitoring has been designed. The radiation-sensitive volumes are four LiF-TLD elements, each covered by its own unique filter. For media-matching, care has been taken to employ nearly tissue equivalent filters of thicknesses of 1000 mg/cm 2 and 300 mg/cm 2 for deep dose and dose to the lens-of-the-eye measurements respectively. Only one metal filter (Cu) is employed to provide low energy photon discrimination. A Thin TL element (0.09 mm thick) is located behind an open window designed to improve the energy under-response to low energy beta rays and to provide closer estimate of the shallow dose equivalent. The deep and shallow dose equivalents are derived from the correlation of the response of the various TL elements to the above quantities through computations based on previously defined relationships obtained from experimental results. The theoretical formalization for the dose calculation algorithms is described in detail, and provides a useful methodology which can be applied to different tissue-equivalent dosimeter assemblies. Experimental data has been obtained by performing irradiation according to the specifications established by DOELAP, using 27 types of pure and mixed radiation fields including Cs-137 gamma rays, low energy photons down to 20 keV, Sr/Y-90, Uranium, and Tl-204 beta particles

  14. Monitoring System for ALICE Surface Areas

    CERN Document Server

    Demirbasci, Oguz

    2016-01-01

    I have been at CERN for 12 weeks within the scope of Summer Student Programme working on a monitoring system project for surface areas of the ALICE experiment during this period of time. The development and implementation of a monitoring system for environmental parameters in the accessible areas where a cheap hardware setup can be deployed were aim of this project. This report explains how it was developed by using Arduino, Raspberry PI, WinCC OA and DIM protocol.

  15. Thermoluminescence dosimeter reader

    International Nuclear Information System (INIS)

    Robertson, M.E.A.; Marshall, J.; Brabants, J.A.P.; Davies, M.E.

    1975-01-01

    An electric circuit arrangement is described including a photomultiplier tube and a high voltage source therefor also includes a feedback loop from the output of the tube to the high voltage source, and loop providing automatic gain stabilization for the tube. The arrangement is used in a dosimeter reader to provide sensitivity correction for the reader each time the reader is to be used

  16. Optical dosimeter

    International Nuclear Information System (INIS)

    Drukaroff, I.; Fishman, R.

    1984-01-01

    A reflecting optical dosimeter is a thin block of optical material having an input light pipe at one corner and an output light pipe at another corner, arranged so that the light path includes several reflections off the edges of the block to thereby greatly extend its length. In a preferred embodiment, one corner of the block is formed at an angle so that after the light is reflected several times between two opposite edges, it is then reflected several more times between the other two edges

  17. Response characteristics of selected personnel neutron dosimeters

    International Nuclear Information System (INIS)

    McDonald, J.C.; Fix, J.J.; Hadley, R.T.; Holbrook, K.L.; Yoder, R.C.; Roberson, P.L.; Endres, G.W.R.; Nichols, L.L.; Schwartz, R.B.

    1983-09-01

    Performance characteristics of selected personnel neutron dosimeters in current use at Department of Energy (DOE) facilities were determined from their evaluation of neutron dose equivalent received after irradiations with specific neutron sources at either the National Bureau of Standards (NBS) or the Pacific Northwest Laboratory (PNL). The characteristics assessed included: lower detection level, energy response, precision and accuracy. It was found that when all of the laboratories employed a common set of calibrations, the overall accuracy was approximately +-20%, which is within uncertainty expected for these dosimeters. For doses above 80 mrem, the accuracy improved to better than 10% when a common calibration was used. Individual differences found in this study may reflect differences in calibration technique rather than differences in the dose rates of actual calibration standards. Second, at dose rates above 100 mrem, the precision for the best participants was generally below +-10% which is also within expected limits for these types of dosimeters. The poorest results had a standard deviation of about +-25%. At the lowest doses, which were sometimes below the lower detection limit, the precision often approached or exceeded +-100%. Third, the lower level of detection for free field 252 Cf neutrons generally ranged between 20 and 50 mrem. Fourth, the energy dependence study provided a characterization of the response of the dosimeters to neutron energies far from the calibration energy. 11 references, 22 figures, 26 tables

  18. Dose mapping of the rectal wall during brachytherapy with an array of scintillation dosimeters

    International Nuclear Information System (INIS)

    Cartwright, L. E.; Suchowerska, N.; Yin, Y.; Lambert, J.; Haque, M.; McKenzie, D. R.

    2010-01-01

    Purpose: In pelvic brachytherapy treatments, the rectum is an organ at risk. The authors have developed an array of scintillation dosimeters suitable for in vivo use that enables quality assurance of the treatment delivery and provides an alert to potential radiation accidents. Ultimately, this will provide evidence to direct treatment planning and dose escalation and correlate dose with the rectal response. Methods: An array of 16 scintillation dosimeters in an insertable applicator has been developed. The dosimeters were calibrated simultaneously in a custom designed circular jig before use. Each dosimeter is optically interfaced to a set of pixels on a CCD camera located outside the treatment bunker. A customized software converts pixel values into dose rate and accumulates dose for presentation during treatment delivery. The performance of the array is tested by simulating brachytherapy treatments in a water phantom. The treatment plans were designed to deliver a known dose distribution on the surface of the rectal applicator, assumed to represent the dose to the rectal wall. Results: The measured doses were compared to those predicted by the treatment plan and found to be in agreement to within the uncertainty in measurement, usually within 3%. The array was also used to track the progression of the source as it moved along the catheter. The measured position was found to agree with the position reported by the afterloader to within the measurement uncertainty, usually within 2 mm. Conclusions: This array is capable of measuring the actual dose received by each region of the rectal wall during brachytherapy treatments. It will provide real time monitoring of treatment delivery and raise an alert to a potential radiation accident. Real time dose mapping in the clinical environment will give the clinician additional confidence to carry out dose escalation to the tumor volume while avoiding rectal side effects.

  19. Evolution of the Regular monitoring of workers using ionizing radiations in medical field in France

    International Nuclear Information System (INIS)

    Biau, A.; Valero, M.; Dubuquoy, E.; Crescini, D.

    2002-01-01

    Among the 260000 workers surveyed by individual monitoring dosimetry in France, about 140000 work in medical field, in Radiodiagnostic. Radiotherapy or Nuclear Medicine. For the twenty last years the global exposure of these workers has decreased continuously like the number of doses over the regular limits of 50 millisieverts per year or even 20 millisieverts which will be included in the new regulations according to the European Directive 96/29 of 13 may 1996. To reinforce this evolution to lower exposures (ALARA) the french regulations have been completed in 1998 and 1999. These new regular prescriptions consist essentially in. - possibility of using another system of passive dosimetry than firm dosimeter (TLD, OSL, RPL...) - obligation of active dosimeter (real time reading) associated with the passive dosimeter for the workers affected in controlled area. - gathering of all the dosimetric results on a data base called SISERI accessible by computer only to the health physicians and the competent person specially authorized. This work is showing from statistical data on the levels of exposure in medical field: - which are the working conditions for which the two system of dosimetry are really necessary - how to delimit the controlled area on realistic basic rather than theoretical assumptions - how the data base SISERI is actually functioning and how it will evaluate from now to 2003. Finally, we propose a schedule to define the different forms of survey for internal and external risks of exposure and to select the best type of monitoring according to the working conditions. (Author)

  20. Evaluation of performance of metal oxide-silicon semiconductor field effect transistor (MOSFET) dosimeter

    International Nuclear Information System (INIS)

    Nagashima, Hiroyuki; Sano, Naoki; Nakamura, Osamu

    2001-01-01

    The JARP level dosimeter is the most suitable for absorbed dose determination in radiotherapy because of its high accuracy. However, in measuring the dose of an extremely small field, a dosimeter with a smaller active region is required. The active region of the MOSFET dosimeter is very small, having a volume of just 0.02 mm 3 . In this study, we evaluated the performance of MOSFET dosimeters with two different sensitivities and examined the usefulness of the MOSFET dosimeter in stereotactic radiosurgery. Using the high-sensitivity MOSFET dosimeter, we were able to reduce the experimental error of absorbed dose (≤±1.8%), and, by correcting the sensitivity, we could use it as a field dosimeter. By turning detectors inside out, we could reduce directional dependence (≤±1.8%). Correction was necessary in the TMR determination because peak depth shifts according to the material of the detector. In the determination of the dose distribution in the penumbra, the resolution of the MOSFET detectors was equal to that of the diamond detector. In the determination of OPF for the extremely small field, better results were obtained with MOSFET than with other small detectors. The high-sensitivity MOSFET dosimeter could properly evaluate the dose of an extremely small field and will be useful in dosimetry of the maximum dose of the field center in stereotactic radiosurgery. (author)

  1. SU-E-T-753: Three-Dimensional Dose Distributions of Incident Proton Particle in the Polymer Gel Dosimeter and the Radiochromic Gel Dosimeter: A Simulation Study with MCNP Code

    International Nuclear Information System (INIS)

    Park, M; Kim, G; Ji, Y; Kim, K; Park, S; Jung, H

    2015-01-01

    Purpose: The purpose of this study is to estimate the three-dimensional dose distributions in the polymer and the radiochromic gel dosimeter, and to identify the detectability of both gel dosimeters by comparing with the water phantom in case of irradiating the proton particles. Methods: The normoxic polymer gel and the LCV micelle radiochromic gel were used in this study. The densities of polymer and the radiochromic gel dosimeter were 1.024 and 1.005 g/cm 3 , respectively. The dose distributions of protons in the polymer and radiochromic gel were simulated using Monte Carlo radiation transport code (MCNPX, Los Alamos National Laboratory). The shape of phantom irradiated by proton particles was a hexahedron with the dimension of 12.4 × 12.4 × 15.0 cm 3 . The energies of proton beam were 50, 80, and 140 MeV energies were directed to top of the surface of phantom. The cross-sectional view of proton dose distribution in both gel dosimeters was estimated with the water phantom and evaluated by the gamma evaluation method. In addition, the absorbed dose(Gy) was also calculated for evaluating the proton detectability. Results: The evaluation results show that dose distributions in both gel dosimeters at intermediated section and Bragg-peak region are similar with that of the water phantom. At entrance section, however, inconsistencies of dose distribution are represented, compared with water. The relative absorbed doses in radiochromic and polymer gel dosimeter were represented to be 0.47 % and 2.26 % difference, respectively. These results show that the radiochromic gel dosimeter was better matched than the water phantom in the absorbed dose evaluation. Conclusion: The polymer and the radiochromic gel dosimeter show similar characteristics in dose distributions for the proton beams at intermediate section and Bragg-peak region. Moreover the calculated absorbed dose in both gel dosimeters represents similar tendency by comparing with that in water phantom

  2. Radiological practices using dosimetric films and electronic dosimeters in F-18 production at cyclotron complex center at IEN/CNEN

    International Nuclear Information System (INIS)

    Santos, Osvaldir P. dos; Silva, J.C.P.; Silva, Luiz Carlos Reina P. da

    2009-01-01

    The aim of this work is to evaluate, guide and correct radiological practices based on dose rate values obtained from dosimetric films and electronic dosimeters used by technicians involved in Fluorine-18 production facilities at IEN/CNEN. Standard statistical methods have been used to analyze and to evaluate these results. The comparison between these results is the first step to orient decisions concerning radiological practices. Besides, radiometric routine surveys are under evaluation in order to improve radiological control in these facility areas. The electronic dosimeters provide the technicians immediate reading and this help them take protective action immediately. So the comparison between dosimetric film and electronic dosimeters data will provide information about how the recently employed electronic dosimeters are being used and what corrections in their practical use are necessary in order to achieve correct practices. In addition the results and observations obtained will be very important to implement possible changes in radiological routine practices in order to optimize them and keep occupationally exposed individuals radiological dose rates, as low as reasonably achievable, according to ALARA principle. (author)

  3. Improvement in the accuracy of polymer gel dosimeters using scintillating fibers

    International Nuclear Information System (INIS)

    Tremblay, Nicolas M; Hubert-Tremblay, Vincent; Bujold, Rachel; Beaulieu, Luc; Lepage, Martin

    2010-01-01

    We propose a novel method for the absolute calibration of polyacrylamide gel (PAG) dosimeters with one or more reference scintillating fiber dosimeters inserted inside the gel. Four calibrated scintillating fibers were inserted into a cylindrical glass container filled with a PAG dosimeter irradiated with a wedge filtered 6 MV photon beam. Calibration curves using small glass vials containing the same gel as the cylindrical containers were used to obtain a first calibration curve. This calibration curve was then adjusted with the dose measured with one of the scintillating fibers in a low gradient part of the field using different approaches. Among these, it was found that a translation of the gel calibration curve yielded the highest accuracy with PAG dosimeters.

  4. Synthesis and production of potassium dichromate as a secondary standard dosimeter; Investigation of its comparative properties with Fricke and clear perspex dosimeters in comprehensive dose mapping of IR-136 Gamma Irradiator

    International Nuclear Information System (INIS)

    Taimoori Seechani; Behzad.

    1995-01-01

    Measurement of the absorbed dose is the principle mode or means of quality control in various application of radiation processing. Specially, the cumulative absorbed dose and its variation in different density product boxes (dose mapping) are often required. Proper discharge of these task would require the use of dosimeters having sufficient degree of accuracy and precision. The secondary standard potassium dichromate dosimeter is synthesized for the first time in Iran. The properties of this dosimeter such as dose response, dose rate dependency, reproducibility, molar extinction coefficient, reduction yield and effect of silver ions on reduction yield of dosimetry solution, irradiation temperature effect as well as pre- and post-irradiation stability of the dosimetry solution have been studied. The comparison of the results of this work with that of the ASTM standard of potassium dichromate has shown very good agreement. The produced dichromate dosimeters were used for comprehensive dose mapping or commissioning of the IR-136 irradiator. The non-uniformity ratio and the isodose lines in various product densities in the IR-136 have been measured. Cumulative dose values were obtained for different rows and levels of the IR-136 irradiator system. The results of dichromate dosimetry, wherever possible, were compared with the response of Fricke and clear perspex dosimeters. At low dose values, the precision of this dosimeter was found to be better than the precision of clear perspex, but about the same as that of Fricke dosimeter. It is thus may be said that this dosimeter has combined the desirable properties of both dosimeter together

  5. Quality assurance of the manual OSLD Badge based personnel monitoring system

    International Nuclear Information System (INIS)

    Kumar, Munish; Gaonkar, U.P.; Koul, D.K.; Datta, D.; Singh, S.K.; Rakesh, R.B.; Kulkarni, M.S.; Ratna, P.

    2018-01-01

    Optically stimulated luminescence (OSL) based technique is widely used for monitoring of radiation workers exposed to x, gamma and beta radiations. Features such as multiple readout, short processing time, better beta response, absence of infra red signal (no heating), flexibility in the dosimeter design using binders or substrates having low melting point are associated advantages. Presently α-Al 2 O 3 :C and BeO based dosimeters are popularly used for personnel monitoring applications worldwide. The present paper describe results of the external quality control check (QAC) tests performed for the present OSLD badge based personnel monitoring system. The results were analyzed as per ANSI criteria and Trumpet curve methods

  6. Is it really not possible to use electronic personal dosimeters in clinical exposure situations?; Sind elektronische Personendosimeter in klinischen Expositionssituationen grundsaetzlich nicht einsetzbar?

    Energy Technology Data Exchange (ETDEWEB)

    Borowski, M. [Klinikum Braunschweig (Germany). Inst. fuer Roentgendiagnostik und Nuklearmedizin; Poppe, B.; Looe, H.K. [Carl von Ossietzky Univ. und Pius-Hospital Oldenburg (Germany). Arbeitsgebiet Medizinische Strahlenphysik; Boetticher, H. von [Klinikum Links der Weser, Bremen (Germany). Inst. fuer Radiologie und Seminar fuer Strahlenschutz

    2010-09-15

    Purpose: Due to significant measuring inaccuracies that can occur under certain conditions, the use of electronic personal dosimeters in statutory measurements in X-ray diagnostics is currently legally restricted. The present study investigates the clinically relevant situations in which measurement errors of more then 20 % can be anticipated. Materials and Methods: Four series of experiments were made, comparing the results of the electronic personal dosimeter EPD Mk2.3 to those of reference dosimeters (TLDs and diagnostic dosimeters). On the one hand, personal doses have been determined in the routine operation of controlled areas in various departments. On the other hand, measurements on phantoms have been conducted in extreme but realistic situations under radiation protection. Experiments were conducted in unweakened scattered radiation as well as in unattenuated and attenuated direct radiation. Results: The tested electronic personal dosimeter type meets the requirements regarding measurement accuracy for 'official' personal dosimeters in all of the examined clinically relevant scattered radiation fields. Only if exposed to radiation directly, an underestimation of the dose can occur and can be greater than 90 %. Conclusion: In the range of scattered radiation of diagnostic X-ray equipment, even in pulsed fields, the use of electronic personal dosimeters is reasonable. Considerable measurement errors can only arise in radiation fields that are not realistic under regular conditions and even in connection with most accidents. (orig.)

  7. X-ray area monitor

    International Nuclear Information System (INIS)

    Nintrakit, N.

    1983-01-01

    The X-ray area monitor is a nuclear electronic device that is essential in radiation protection in high radiation laboratories, e.g. in medical diagnosis using X-rays and in industrial X-radiography. Accidentally the level of X-radiator may arise above the safe permissible level and in such a case the alarm system of the area monitor will work and disconnect the ac power supply form the X-ray unit. Principally the device is a radiation counter using G.M.tube as radiation detector with high voltage supply variable form 200 to 2,000 volts. The maximum count rate of the scaler is 1.5 MHz and the total count is displayed on 4 digit LED's. A time base is used to control the counting time, the frequency multiplier, radiation safety limit, comparator and the radiation hazard warning signal. The reliability of the instrument is further enhanced through the addition of the random correction circuit, and it is applicable both in X- and γ -radiation

  8. Development and characterization of a three-dimensional radiochromic film stack dosimeter for megavoltage photon beam dosimetry.

    Science.gov (United States)

    McCaw, Travis J; Micka, John A; DeWerd, Larry A

    2014-05-01

    Three-dimensional (3D) dosimeters are particularly useful for verifying the commissioning of treatment planning and delivery systems, especially with the ever-increasing implementation of complex and conformal radiotherapy techniques such as volumetric modulated arc therapy. However, currently available 3D dosimeters require extensive experience to prepare and analyze, and are subject to large measurement uncertainties. This work aims to provide a more readily implementable 3D dosimeter with the development and characterization of a radiochromic film stack dosimeter for megavoltage photon beam dosimetry. A film stack dosimeter was developed using Gafchromic(®) EBT2 films. The dosimeter consists of 22 films separated by 1 mm-thick spacers. A Virtual Water™ phantom was created that maintains the radial film alignment within a maximum uncertainty of 0.3 mm. The film stack dosimeter was characterized using simulations and measurements of 6 MV fields. The absorbed-dose energy dependence and orientation dependence of the film stack dosimeter were investigated using Monte Carlo simulations. The water equivalence of the dosimeter was determined by comparing percentage-depth-dose (PDD) profiles measured with the film stack dosimeter and simulated using Monte Carlo methods. Film stack dosimeter measurements were verified with thermoluminescent dosimeter (TLD) microcube measurements. The film stack dosimeter was also used to verify the delivery of an intensity-modulated radiation therapy (IMRT) procedure. The absorbed-dose energy response of EBT2 film differs less than 1.5% between the calibration and film stack dosimeter geometries for a 6 MV spectrum. Over a series of beam angles ranging from normal incidence to parallel incidence, the overall variation in the response of the film stack dosimeter is within a range of 2.5%. Relative to the response to a normally incident beam, the film stack dosimeter exhibits a 1% under-response when the beam axis is parallel to the film

  9. Temperature, humidity and time., Combined effects on radiochromic film dosimeters

    International Nuclear Information System (INIS)

    Abdel-Fattah, A.A.; Miller, A.

    1996-01-01

    The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Riso B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60 o C for irradiation by 60 Co photons and 10-MeV electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is + 0.25 ± 0.1% per o C for the FWT-60-00 dosimeters and +0.5 ± 0.1% per o C for Riso B3 dosimeters at temperatures between 20 and 50 o C and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger dependences. Whenever possible one should use dosimeters sealed in pouches under controlled intermediate humidity conditions (30-50%) or, if that is impractical, one should maintain conditions of calibration as close as possible to the conditions of use. Without that precaution, severe dosimetry errors may result. (author)

  10. Characteristics Of Dosimeter TL CaSO4:Dy Glass Capillaries For Environmental Radiation Dose Monitoring

    International Nuclear Information System (INIS)

    Sofyan Hasnel; Yulianti, Helfi; Ramain, Abubakar; Kusumawati, Dyah D.; Suyati

    2000-01-01

    research on the characteristic of dosimeter TL CaSO 4 : Dy glass capillaries for environmental dose radiation have been carried out. The results obtained are uniform response and reproducibility during three cycles consumption with average percentage standard deviation of 7.31 % and 5.45%. The response dose is linear and has a minimum detectable dose of 0.01 mGy, sunshine effect with non-penetrating light capsule of 4.65%, humidity effects is not significant by using non-penetrating light capsule. Radiation dose information during 30 days are fading 25%

  11. Description and evaluation of the Hanford personnel dosimeter program from 1944 through 1989. [Contain Glossary

    Energy Technology Data Exchange (ETDEWEB)

    Wilson, R.H.; Fix, J.J.; Baumgartner, W.V.; Nichols, L.L.

    1990-09-01

    This report describes the evolution of personnel dosimeter technology at Hanford since the inception of Hanford operations in 1944. Each of the personnel dosimeter systems used by people working or visiting Hanford is described. In addition, the procedures used to calibrate and calculate dose for each of the dosimeter systems are described. The accuracy of the recorded dose, primarily whole body deep dose, for the different dosimeter systems is evaluated. The evaluation is based on an extensive review of historical literature, as well as a 1989 intercomparison study of all film dosimeters and performance testing of the thermoluminescent dosimeter, also conducted during 1989. 73 refs., 40 figs., 41 tabs.

  12. Characterization of MOSFET Dosimeter Angular Response Using a Spherical Phantom for Fluoroscopic Dosimetry.

    Science.gov (United States)

    Wang, Chu; Hill, Kevin; Yoshizumi, Terry

    2016-01-01

    Metal-Oxide Semiconductor Field-Effect Transistor (MOSFET) dosimeters, placed in anthropomorphic phantoms, are a standard method for organ dosimetry in medical x-ray imaging applications. However, many x-ray applications, particularly fluoroscopy procedures, use variable projection angles. During dosimetry, the MOSFET detector active area may not always be perpendicular to the x-ray beam. The goal of this study was to characterize the dosimeter's angular response in the fluoroscopic irradiation involved in pediatric cardiac catheterization procedures, during which a considerable amount of fluoroscopic x-ray irradiation is often applied from various projection angles. A biological x-ray irradiator was used to simulate the beam quality of a biplane fluoroscopy imaging system. A custom-designed acrylic spherical scatter phantom was fabricated to measure dosimeter response (in mV) in two rotational axes, axial (ψ) and normal-to-axial (θ), in 30° increments, as well as four common oblique angles used in cardiac catheterization: a) 90° Left Anterior Oblique (LAO); b) 70° LAO/ 20° Cranial; c) 20° LAO/ 15° Cranial; and d) 30° Right Anterior Oblique (RAO). All results were normalized to the angle where the dosimeter epoxy is perpendicular to the beam or the Posterior-Anterior projection angle in the clinical setup. The relative response in the axial rotation was isotropic (within ± 10% deviation); that in the normal-to-axial rotation was isotropic in all angles except the ψ = 270° angle, where the relative response was 83 ± 9%. No significant deviation in detector response was observed in the four common oblique angles, with their relative responses being: a) 102 ± 3%; b) 90 ± 3%; c) 92 ± 3%; and d) 95 ± 3%, respectively. These angular correction factors will be used in future dosimetry studies for fluoroscopy. The spherical phantom may be useful for other applications, as it allows the measurement of dosimeter response in virtually all angles in the 3

  13. Performances of Dose Measurement of Commercial Electronic Dosimeters using Geiger Muller Tube and PIN Diode

    Energy Technology Data Exchange (ETDEWEB)

    Yoo, Hyunjun; Kim, Chankyu; Kim, Yewon; Kim, Giyoon; Cho, Gyuseong [Korea Advanced Institute of Science and Technology, Daejeon (Korea, Republic of)

    2014-05-15

    There are two categories in personal dosimeters, one is passive type dosimeter such as TLD (thermoluminescence dosimeter) and the other is active type dosimeter such as electronic dosimeter can show radiation dose immediately while TLD needs long time to readout its data by heating process. For improving the reliability of measuring dose for any energy of radiations, electronic dosimeter uses energy filter by metal packaging its detector using aluminum or copper, but measured dose of electronic dosimeter with energy filter cannot be completely compensated in wide radiation energy region. So, in this paper, we confirmed the accuracy of dose measurement of two types of commercial EPDs using Geiger Muller tube and PIN diode with CsI(Tl) scintillator in three different energy of radiation field. The experiment results for Cs-137 was almost similar with calculation value in the results of both electronic dosimeters, but, the other experiment values with Na-22 and Co-60 had higher error comparing with Cs-137. These results were caused by optimization of their energy filters. The optimization was depending on its thickness of energy filter. So, the electronic dosimeters have to optimizing the energy filter for increasing the accuracy of dose measurement or the electronic dosimeter using PIN diode with CsI(Tl) scintillator uses the multi-channel discriminator for using its energy information.

  14. LLL development of a combined etch track: albedo dosimeter

    International Nuclear Information System (INIS)

    Griffith, R.V.; Fisher, J.C.; Harder, C.A.

    1977-01-01

    The addition of polycarbonate sheet to albedo detectors for electrochemical etching provides a simple, inexpensive way to reduce the spectral sensitivity of the personnel dosimeter without losing the albedo features of sensitivity and ease of automation. The ECEP technique also provides the dosimetrist with the potential for identifying conditions of body orientation that might otherwise lead to significant error in dosimeter evaluation

  15. Temperature, humidity and time. Combined effects on radiochromic film dosimeters

    DEFF Research Database (Denmark)

    Abdel-Fattah, A.A.; Miller, A.

    1996-01-01

    The effects of both relative humidity and temperature during irradiation on the dose response of FWT-60-00 and Riso B3 radiochromic film dosimeters have been investigated in the relative humidity (RH) range 11-94% and temperature range 20-60 degrees C for irradiation by Co-60 photons and 10-Me......V electrons. The results show that humidity and temperature cannot be treated as independent variables, rather there appears to be interdependence between absorbed dose, temperature, and humidity. Dose rate does not seem to play a significant role. The dependence of temperature during irradiation is +0.......25 +/- 0.1% per degrees C for the FWT-60-00 dosimeters and +0.5 +/- 0.1% per degrees C For Riso B3 dosimeters at temperatures between 20 and 50 degrees C and at relative humidities between 20 and 53%. At extreme conditions both with respect to temperature and to humidity, the dosimeters show much stronger...

  16. Technical Evaluation of Radiation Dose Delivered in Prostate Cancer Patients as Measured by an Implantable MOSFET Dosimeter

    International Nuclear Information System (INIS)

    Beyer, Gloria P.; Scarantino, Charles W.; Prestidge, Bradley R.; Sadeghi, Amir G.; Anscher, Mitchell S.; Miften, Moyed; Carrea, Tammy B.; Sims, Marianne C.; Black, Robert D.

    2007-01-01

    Purpose: To perform a comparison of the daily measured dose at depth in tissue with the predicted dose values from treatment plans for 29 prostate cancer patients involved in a clinical trial. Methods and Materials: Patients from three clinical sites were implanted with one or two dosimeters in or near the prostatic capsule. The implantable device, known as the DVS, is based on a metal-oxide-semiconductor field effect transistor (MOSFET) detector. A portable telemetric readout system couples to the dosimeter antenna (visible on kilovoltage, computed tomography, and ultrasonography) for data transfer. The predicted dose values were determined by the location of the MOSFET on the treatment planning computed tomography scan. Serial computed tomography images were taken every 2 weeks to evaluate any migration of the device. The clinical protocol did not permit alteration of the treatment parameters using the dosimeter readings. For some patients, one of several image-guided radiotherapy (RT) modalities was used for target localization. Results: The evaluation of dose discrepancy showed that in many patients the standard deviation exceeded the previous values obtained for the dosimeter in a phantom. In some patients, the cumulative dose disagreed with the planned dose by ≥5%. The data presented suggest that an implantable dosimeter can help identify dose discrepancies (random or systematic) for patients treated with external beam RT and could be used as a daily treatment verification tool for image-guided RT and adaptive RT. Conclusion: The results of our study have shown that knowledge of the dose delivered per fraction can potentially prevent over- or under-dosage to the treatment area and increase the accuracy of RT. The implantable dosimeter could also be used as a localizer for image-guided RT

  17. Passive dosimetry: introduction of a new dosimeter based on OSL technology

    International Nuclear Information System (INIS)

    Archambault, V.; Le Roy, G.; Prugnaud, B.

    2005-01-01

    A new passive dosimeter based on OSL technology has been introduced on the French market. In this article are described: the technology and the material on which this new detector relied, the dosimeter itself. (author)

  18. Assessment of radiation exposure in dental cone-beam computerized tomography with the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters and Monte Carlo simulations.

    Science.gov (United States)

    Koivisto, J; Kiljunen, T; Tapiovaara, M; Wolff, J; Kortesniemi, M

    2012-09-01

    The aims of this study were to assess the organ and effective dose (International Commission on Radiological Protection (ICRP) 103) resulting from dental cone-beam computerized tomography (CBCT) imaging using a novel metal-oxide semiconductor field-effect transistor (MOSFET) dosimeter device, and to assess the reliability of the MOSFET measurements by comparing the results with Monte Carlo PCXMC simulations. Organ dose measurements were performed using 20 MOSFET dosimeters that were embedded in the 8 most radiosensitive organs in the maxillofacial and neck area. The dose-area product (DAP) values attained from CBCT scans were used for PCXMC simulations. The acquired MOSFET doses were then compared with the Monte Carlo simulations. The effective dose measurements using MOSFET dosimeters yielded, using 0.5-cm steps, a value of 153 μSv and the PCXMC simulations resulted in a value of 136 μSv. The MOSFET dosimeters placed in a head phantom gave results similar to Monte Carlo simulations. Minor vertical changes in the positioning of the phantom had a substantial affect on the overall effective dose. Therefore, the MOSFET dosimeters constitute a feasible method for dose assessment of CBCT units in the maxillofacial region. Copyright © 2012 Elsevier Inc. All rights reserved.

  19. Individual monitoring

    International Nuclear Information System (INIS)

    2004-01-01

    This Practical Radiation Technical Manual is one of a series which has been designed to provide guidance on radiological protection for employers, Radiation Protection Officers, managers and other technically competent persons who have a responsibility to ensure the safety of employees working with ionizing radiation. The Manual may be used together with the appropriate IAEA Practical Radiation Safety Manual to provide adequate training, instruction or information on individual monitoring for all employees engaged in work with ionizing radiations. Sources of ionizing radiation have a large number of applications in the workplace. The exposures of the individual workers involved may need to be routinely monitored and records kept of their cumulative radiation doses. There are also occasions when it is necessary to retrospectively determine a dose which may have been received by a worker. This Manual explains the basic terminology associated with individual monitoring and describes the principal types of dosimeters and other related techniques and their application in the workplace. The Manual will be of most benefit if it forms part of more comprehensive training or is supplemented by the advice of a qualified expert in radiation protection. Most of the dosimeters and techniques described in this Manual can only be provided by qualified experts

  20. MO-AB-BRA-04: Radiation Measurements with a DNA Double-Strand-Break Dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Obeidat, M; Cline, K; Stathakis, S; Papanikolaou, N; Rasmussen, K; Gutierrez, A; Ha, CS; Lee, SE; Shim, EY; Kirby, N [University of Texas HSC SA, San Antonio, TX (United States)

    2016-06-15

    Purpose: Many types of dosimeters are used to measure radiation, but none of them directly measures the biological effect of this dose. The purpose here is to create a dosimeter that can measure the probability of double-strand breaks (DSB) for DNA, which is directly related to the biological effect of radiation. Methods: The dosimeter has DNA strands, which are labeled on one end with biotin and on the other with fluorescein. The biotin attaches these strands to magnetic beads. We suspended the DNA dosimeter in phosphate-buffered saline (PBS) as it matches the internal environment of the body. We placed small volumes (50µL) of the DNA dosimeter into tubes and irradiated these samples in a water-equivalent plastic phantom with several doses (three samples per dose). After irradiating the samples, a magnet was placed against the tubes. The fluorescein attached to broken DNA strands was extracted (called the supernatant) and placed into a different tube. The fluorescein on the unbroken strands remained attached to the beads in the tube and was re-suspended with 50µL of PBS. A fluorescence reader was used to measure the fluorescence for both the re-suspended beads and supernatant. To prove that we are measuring DSB, we tested dosimeter response with two different lengths of attached DNA strands (1 and 4 kilo-base pair). Results: The probability of DSB at the dose levels of 5, 10, 25, and 50 Gy were 0.05, 0.08, 0.12, and 0.19, respectively, while the coefficients of variation were 0.14, 0.07, 0.02, and 0.01, respectively. The 4 kilo-base-pair dosimeter produced 5.3 times the response of the 1 kilo-base-pair dosimeter. Conclusion: The DNA dosimeter yields a measurable response to dose that scales with the DNA strand length. The goal now is to refine the dosimeter fabrication to reproducibly create a low coefficient of variation for the lower doses. This work was supported in part by Yarmouk University (Irbid, Jordan) and CPRIT (RP140105)

  1. MO-AB-BRA-04: Radiation Measurements with a DNA Double-Strand-Break Dosimeter

    International Nuclear Information System (INIS)

    Obeidat, M; Cline, K; Stathakis, S; Papanikolaou, N; Rasmussen, K; Gutierrez, A; Ha, CS; Lee, SE; Shim, EY; Kirby, N

    2016-01-01

    Purpose: Many types of dosimeters are used to measure radiation, but none of them directly measures the biological effect of this dose. The purpose here is to create a dosimeter that can measure the probability of double-strand breaks (DSB) for DNA, which is directly related to the biological effect of radiation. Methods: The dosimeter has DNA strands, which are labeled on one end with biotin and on the other with fluorescein. The biotin attaches these strands to magnetic beads. We suspended the DNA dosimeter in phosphate-buffered saline (PBS) as it matches the internal environment of the body. We placed small volumes (50µL) of the DNA dosimeter into tubes and irradiated these samples in a water-equivalent plastic phantom with several doses (three samples per dose). After irradiating the samples, a magnet was placed against the tubes. The fluorescein attached to broken DNA strands was extracted (called the supernatant) and placed into a different tube. The fluorescein on the unbroken strands remained attached to the beads in the tube and was re-suspended with 50µL of PBS. A fluorescence reader was used to measure the fluorescence for both the re-suspended beads and supernatant. To prove that we are measuring DSB, we tested dosimeter response with two different lengths of attached DNA strands (1 and 4 kilo-base pair). Results: The probability of DSB at the dose levels of 5, 10, 25, and 50 Gy were 0.05, 0.08, 0.12, and 0.19, respectively, while the coefficients of variation were 0.14, 0.07, 0.02, and 0.01, respectively. The 4 kilo-base-pair dosimeter produced 5.3 times the response of the 1 kilo-base-pair dosimeter. Conclusion: The DNA dosimeter yields a measurable response to dose that scales with the DNA strand length. The goal now is to refine the dosimeter fabrication to reproducibly create a low coefficient of variation for the lower doses. This work was supported in part by Yarmouk University (Irbid, Jordan) and CPRIT (RP140105)

  2. Design and development of a PMOSFET gamma ray dosimeter

    International Nuclear Information System (INIS)

    Khanna, V.K.; Kumar, A.; Gupta, R.P.; Pandya, A.; Roy, Rajesh

    2005-01-01

    A P-channel MOSFET chip has been designed for detection of gamma radiations. The chip consists of three MOSFETs of different geometrical parameters for achieving sensitivity to low and high dose ranges. One of the MOSFET structures has a closed geometry to reduce the leakage current. The developed dosimeter being a MOSFET, its IC (Integrated Circuit)-compatibility helps in easy interfacing with readout circuitry. The dosimeter fabrication process is based on metal-gate MOSFET technology with thick gate oxide to increase the effective number of electron-hole pairs generated by the gamma rays impinging on the device. The process for the chip realization has been designed and simulated to achieve the required impurity diffusion profile. The chip has been fabricated using the above process and electrically characterized. The device has been exposed to gamma ray source and its characteristics measured. The change in threshold voltage of the MOSFET after exposure has been used to calculate the sensitivity of the device. The developed dosimeter has potential applications in personnel dosimetry and cancer treatment. This paper describes the basic detection mechanism of the MOSFET, the design approach, and fabrication process of the MOSFET dosimeter. (author)

  3. Angular dependence of response of dosimeters exposed to an extended radioactive source

    International Nuclear Information System (INIS)

    Manai, K.; Trabelsi, A.; Madouri, F.

    2014-01-01

    This study was carried out to investigate the exposure angular dependence of dosimeters response when exposed to the extended gamma source of an irradiation facility. Using analytical and Monte Carlo analysis, we show that dosimeters response has no angular dependence as claimed by a previous study. The dose rate formula we derived takes into account the path length of the photons in the dosimeter. Experimental data have been used to validate our analytical and Monte Carlo methods. Furthermore, the effects on the dosimeters responses in relation to their sizes response of their size and geometry and orientation have been investigated and, within statistical errors, no angular dependence was found. - Highlights: • We investigate the exposer angle dependence of dosimeter response to a gamma source. • Analytical and Monte Carlo analyses show no angular dependence as claimed by others. • We derive the dose rate formulae taking into account the path length of photons. • Analytical and Monte Carlo models have been validated using experimental data

  4. Water equivalence of NIPAM based polymer gel dosimeters with enhanced sensitivity for x-ray CT

    Science.gov (United States)

    Gorjiara, Tina; Hill, Robin; Bosi, Stephen; Kuncic, Zdenka; Baldock, Clive

    2013-10-01

    Two new formulations of N-isopropylacrylamide (NIPAM) based three dimensional (3D) gel dosimeters have recently been developed with improved sensitivity to x-ray CT readout, one without any co-solvent and the other one with isopropanol co-solvent. The water equivalence of the NIPAM gel dosimeters was investigated using different methods to calculate their radiological properties including: density, electron density, number of electrons per grams, effective atomic number, photon interaction probabilities, mass attenuation and energy absorption coefficients, electron collisional, radiative and total mass stopping powers and electron mass scattering power. Monte Carlo modelling was also used to compare the dose response of these gel dosimeters with water for kilovoltage and megavoltage x-ray beams and for megavoltage electron beams. We found that the density and electron density of the co-solvent free gel dosimeter are more water equivalent with less than a 2.6% difference compared to a 5.7% difference for the isopropanol gel dosimeter. Both the co-solvent free and isopropanol solvent gel dosimeters have lower effective atomic numbers than water, differing by 2.2% and 6.5%, respectively. As a result, their photoelectric absorption interaction probabilities are up to 6% and 19% different from water, respectively. Compton scattering and pair production interaction probabilities of NIPAM gel with isopropanol differ by up to 10% from water while for the co-solvent free gel, the differences are 3%. Mass attenuation and energy absorption coefficients of the co-solvent free gel dosimeter and the isopropanol gel dosimeter are up to 7% and 19% lower than water, respectively. Collisional and total mass stopping powers of both gel dosimeters differ by less than 2% from those of water. The dose response of the co-solvent free gel dosimeter is water equivalent (with x-ray beams over the energy range 180 keV-18 MV, both gel dosimeters have less than 2% discrepancy with water. For

  5. Design of movable fixed area γ dose rate monitor

    International Nuclear Information System (INIS)

    Li Dongyu; Cheng Wen; Li Jikai; Huang Hong; Shen Qiming; Zhang Qiang; Liu Zhengshan

    2005-10-01

    Movable fixed area γ dose rate monitor has not only the characteristics of fixed area γ dose rate monitor, but that of portable meter as well. Its main function is to monitor the areas where dose rate would change without orderliness to prevent unplanned radiation exposure accidents from happening. The design way of the monitor, the main indicators description, the working principle and the comprising of software and hardware are briefly introduced. The monitor has the characteristics of simple installation, easy maintenance, little power consumption, wide range, notability of visual and audible alarm and so on. Its design and technique have novelty and advancement. (authors)

  6. Scintillation counter based radiation dosimeter

    International Nuclear Information System (INIS)

    Shin, Jeong Hyun

    2009-02-01

    environmental radiation. The operation mechanism is quite simple. First step is scintillation in scintillator. Second step is conversion from photon to charge in photodiode. Third step is signal amplification in charge amplifier. Fourth step is pulse shaping for discrimination in shaping amp. Fifth step is reduction of noise pulse in discriminator. Final step is pulse counting in counter. This new radiation detection system can give reliable radiation dose data as low cost. So we can supply this new radiation dosimeter to the area near radiation and nuclear facilities. I will hope this new radiation dosimeter can contribute to the public acceptance on radiation use and nuclear facilities

  7. A critical assessment of two types of personal UV dosimeters.

    Science.gov (United States)

    Seckmeyer, Gunther; Klingebiel, Marcus; Riechelmann, Stefan; Lohse, Insa; McKenzie, Richard L; Liley, J Ben; Allen, Martin W; Siani, Anna-Maria; Casale, Giuseppe R

    2012-01-01

    Doses of erythemally weighted irradiances derived from polysulphone (PS) and electronic ultraviolet (EUV) dosimeters have been compared with measurements obtained using a reference spectroradiometer. PS dosimeters showed mean absolute deviations of 26% with a maximum deviation of 44%, the calibrated EUV dosimeters showed mean absolute deviations of 15% (maximum 33%) around noon during several test days in the northern hemisphere autumn. In the case of EUV dosimeters, measurements with various cut-off filters showed that part of the deviation from the CIE erythema action spectrum was due to a small, but significant sensitivity to visible radiation that varies between devices and which may be avoided by careful preselection. Usually the method of calibrating UV sensors by direct comparison to a reference instrument leads to reliable results. However, in some circumstances the quality of measurements made with simple sensors may be over-estimated. In the extreme case, a simple pyranometer can be used as a UV instrument, providing acceptable results for cloudless skies, but very poor results under cloudy conditions. It is concluded that while UV dosimeters are useful for their design purpose, namely to estimate personal UV exposures, they should not be regarded as an inexpensive replacement for meteorological grade instruments. © 2011 Wiley Periodicals, Inc. Photochemistry and Photobiology © 2011 The American Society of Photobiology.

  8. PNNL Results from 2010 CALIBAN Criticality Accident Dosimeter Intercomparison Exercise

    International Nuclear Information System (INIS)

    Hill, Robin L.; Conrady, Matthew M.

    2011-01-01

    This document reports the results of the Hanford personnel nuclear accident dosimeter (PNAD) and fixed nuclear accident dosimeter (FNAD) during a criticality accident dosimeter intercomparison exercise at the CEA Valduc Center on September 20-23, 2010. Pacific Northwest National Laboratory (PNNL) participated in a criticality accident dosimeter intercomparison exercise at the Commissariat a Energie Atomique (CEA) Valduc Center near Dijon, France on September 20-23, 2010. The intercomparison exercise was funded by the U.S. Department of Energy, Nuclear Criticality Safety Program, with Lawrence Livermore National Laboratory as the lead Laboratory. PNNL was one of six invited DOE Laboratory participants. The other participating Laboratories were: Lawrence Livermore National Laboratory (LLNL), Los Alamos National Laboratory (LANL), Savannah River Site (SRS), the Y-12 National Security Complex at Oak Ridge, and Sandia National Laboratory (SNL). The goals of PNNL's participation in the intercomparison exercise were to test and validate the procedures and algorithm currently used for the Hanford personnel nuclear accident dosimeters (PNADs) on the metallic reactor, CALIBAN, to test exposures to PNADs from the side and from behind a phantom, and to test PNADs that were taken from a historical batch of Hanford PNADs that had varying degrees of degradation of the bare indium foil. Similar testing of the PNADs was done on the Valduc SILENE test reactor in 2009 (Hill and Conrady, 2010). The CALIBAN results are reported here.

  9. Calibration of film dosimeters by means of absorbed dose calorimeters

    International Nuclear Information System (INIS)

    Nikolaev, S.M.; Vanyushkin, B.M.; Kon'kov, N.G.

    1980-01-01

    Methods of graduating film dosimeters by means of calorimeters of absorbed doses, are considered. Graduating of film dosimeters at the energies of accelerated electrons from 4 to 10 MeV can be carried out by means of quasiadiabatic calorimeter of local absorption, the absorber thickness of which should not exceed 5-10% of Rsub(e) value, where Rsub(e) - free electron path of the given energy. In this case film is located inside the calorimeter. For graduating films with thickness not less than (0.1-0.2)Rsub(e) it is suggested to use calorimeter of full absorption; then the graduated dosimeters are located in front of the calorimeter. Graduation of films at small energies of electrons is exercised by means of a package of films, approximately Rsub(e) thick. A design of quasiadiabatic calorimeter, intended for graduating dosimeters within the energy range of electron beam from 4 to 10 MeV, is considered. The quasiadiabatic calorimeter is a thin graphite tablet with heater and thermocouple, surrounded by foam plastic thermostating case. Electricity quantity, accumulated during the radiation field pass, is measured in the case of using the quasiadiabatic calorimeter for film graduating. The results of graduating film dosimeters, obtained using film package with Rsub(e) thickness, are presented. The obtained results coincide within 5% limits with the data known beforehand [ru

  10. Characterization of lithium formate EPR dosimeters for high dose applications – comparison with alanine

    DEFF Research Database (Denmark)

    Waldeland, Einar; Helt-Hansen, Jakob; Malinen, Eirik

    2011-01-01

    Lithium formate and l-α-alanine (alanine) EPR dosimeters were irradiated to doses from 100 Gy to 100 kGy. The irradiations were mainly performed at a Gammacell irradiator with dose rate of approximately 5.5 kGy h−1. Both the peak-to-peak amplitude of the first derivative EPR spectrum and the area...... irradiated to 1 kGy at temperatures from 11ºC to 40ºC were analyzed. By fitting an ‘exponential rise to maximum'-function to the dependence of the area under the EPR absorption spectrum on the dose, saturation doses of 53 kGy and 87 kGy for lithium formate and alanine, respectively, were found. Lower...... estimates were found when analyzing the dose dependence of the peak-to-peak amplitude. Furthermore, the peak-to-peak width was found to increase for doses above 10 kGy. No dose rate dependence for any of the studied materials was observed and the temperature coefficients at 25ºC (i.e. change in dosimeter...

  11. IAEA reference dosimeter: Alanine-ESR

    International Nuclear Information System (INIS)

    Mehta, K.; Girzikowsky, R.

    1999-01-01

    Since 1985, the IAEA has been using alanine-ESR as a transfer dosimeter for its dose quality audit service, namely the International Dose Assurance Service. The alanine dosimeters are rod-type containing 70 wt% DL--α-alanine and 30 wt% polystyrene. We have two self-shielded gamma facilities for the calibration of the dosimetry system, where the temperature within the irradiation chamber can be controlled by a specially designed unit. A 4th order polynomial is fitted to the 16 data points in the dose range of 100 Gy to 50 kGy. The measured value of the irradiation temperature coefficient at two dose values (15 and 45 kGy) is 0.23 %/deg. C. Also, the ESR-response was followed for several dosimeters for about 8 months to study the post-irradiation effect. A value of 0.008 %/day was observed for the fading of the response for two dose values (15 and 45 kGy) and three irradiation temperatures (15, 27 and 40 deg. C). The effect of the analysis temperature on the ESR response was also studied. The combined relative uncertainty for the IAEA alanine-ESR dosimetry system is 1.5% (k=1). This includes that transferred from the primary laboratory for the dose rate measurements of the gamma facilities, dosimetry system calibration uncertainties, batch variability and uncertainty in the curve fitting procedure. This value however does not include the contribution due to the irradiation temperature correction which is applied when it differs from that during calibration; this component being specific for each dose measurement. (author)

  12. Methacholine challenge test: Comparison of tidal breathing and dosimeter methods in children.

    Science.gov (United States)

    Mazi, Ahlam; Lands, Larry C; Zielinski, David

    2018-02-01

    Methacholine Challenge Test (MCT) is used to confirm, assess the severity and/or rule out asthma. Two MCT methods are described as equivalent by the American Thoracic Society (ATS), the tidal breathing and the dosimeter methods. However, the majority of adult studies suggest that individuals with asthma do not react at the same PC 20 between the two methods. Additionally, the nebulizers used are no longer available and studies suggest current nebulizers are not equivalent to these. Our study investigates the difference in positive MCT tests between three methods in a pediatric population. A retrospective, chart review of all MCT performed with spirometry at the Montreal Children's Hospital from January 2006 to March 2016. A comparison of the percentage positive MCT tests with three methods, tidal breathing, APS dosimeter and dose adjusted DA-dosimeter, was performed at different cutoff points up to 8 mg/mL. A total of 747 subjects performed the tidal breathing method, 920 subjects the APS dosimeter method, and 200 subjects the DA-dosimeter method. At a PC 20 cutoff ≤4 mg/mL, the percentage positive MCT was significantly higher using the tidal breathing method (76.3%) compared to the APS dosimeter (45.1%) and DA-dosimeter (65%) methods (P < 0.0001). The choice of nebulizer and technique significantly impacts the rate of positivity when using MCT to diagnose and assess asthma. Lack of direct comparison of techniques within the same individuals and clinical assessment should be addressed in future studies to standardize MCT methodology in children. © 2017 Wiley Periodicals, Inc.

  13. NRC TLD Direct Radiation Monitoring Network. Progress report, January-June 1981

    International Nuclear Information System (INIS)

    1982-04-01

    This report provides the status and results of the NRC Thermoluminescent Dosimeter (TLD) Direct Radiation Monitoring Network. It presents the radiation levels measured in the vicinity of 55 NRC-licensed facility sites throughout the country for the first half of 1981. The program objectives, scope, and methodology are given. The TLD system, dosimeter location, data processing scheme, and quality assurance program are outlined

  14. CAP vessel monitoring. Programme, measurement and neutron calculation

    International Nuclear Information System (INIS)

    Farrugia, J.M.; Nimal, J.C.; Totth, B.; Lloret, R.; Perdreau, R.

    1982-03-01

    Starting with the design of the CAP (Prototype Advanced NSSS), a programme for pressure vessel monitoring has been prepared, including dosimetry. The dosimetry programme encompasses activation dosimeters (Cu, Nb, Co) and fission dosimeters ( 237 Np, 238 U) installed either inside the pressure vessel with the monitoring test-samples, or in a counting tube outside the pressure vessel. In the first place, a description of the method for neutronic calculation is given; such calculations use the codes ANISN and MERCURE 4 allowing assessment of the neutron spectrum seen by the detectors and the related reaction coefficient. This is followed by a description of the instrumentation. The initial dosimetry results available after the initial operating cycles concur with calculations [fr

  15. Establishing personal dosimetry procedure using optically stimulated luminescence dosimeters in photon and mixed photon-neutron radiation fields

    International Nuclear Information System (INIS)

    Le Ngoc Thiem; Bui Duc Ky; Trinh Van Giap; Nguyen Huu Quyet; Ho Quang Tuan; Vu Manh Khoi; Chu Vu Long

    2017-01-01

    According to Vietnamese Law on Atomic Energy, personal dosimetry (PD) for radiation workers is required periodically in order to fulfil the national legal requirements on occupational radiation dose management. Since the radiation applications have become popular in Vietnamese society, the thermal luminescence dosimeters (TLDs) have been used as passive dosimeters for occupational monitoring in the nation. Together with the quick increase in radiation applications and the number of personnel working in radiation fields, the Optically Stimulated Luminescence Dosimeters (OSLDs) have been first introduced since 2015. This work presents the establishment of PD measuring procedure using OSLDs which are used for measuring photons and betas known as Inlight model 2 OSL (OSLDs-p,e) and for measuring mixed radiations of neutrons, photons and betas known as Inlight LDR model 2 (OSLDs-n,p,e). Such following features of OSLDs are investigated: detection limit, energy response, linearity, reproducibility, angular dependency and fading with both types of OSLDs-p,e and OSLDs-n,p,e. The result of an intercomparison in PD using OSLDs is also presented in the work. The research work also indicates that OSL dosimetry can be an alternative method applied in PD and possibly become one of the most popular personal dosimetry method in the future. (author)

  16. TH-CD-201-11: Optimizing the Response and Cost of a DNA Double-Strand Break Dosimeter

    Energy Technology Data Exchange (ETDEWEB)

    Obeidat, M; Cline, K; Stathakis, S; Papanikolaou, N; Rasmussen, K; Gutierrez, A; Ha, CS; Lee, SE; Shim, EY; Kirby, N [University of Texas HSC SA, San Antonio, TX (United States)

    2016-06-15

    Purpose: A DNA double-strand break (DSB) dosimeter was developed to measure the biological effect of radiation. The goal here is to refine the fabrication method of this dosimeter to reproducibly create a low coefficient of variation (CoV) and reduce the cost for the dosimeter. Methods: Our dosimeter consists of 4 kilo-base pair DNA strands (labeled on one end with biotin and on the other with fluorescein) attached to streptavidin magnetic beads. The final step of the DNA dosimeter fabrication is to suspend these attached beads in phosphate-buffered saline (PBS). The amount of PBS used to suspend the attached beads and the relative volume of the DNA strands to the beads both affect the CoV and dosimeter cost. We diluted the beads attached with DNA in different volumes of PBS (100, 200, and 400 µL) to create different concentrations of the DNA dosimeter. Then we irradiated these dosimeters (50 µL samples) in a water-equivalent plastic phantom at 25 and 50 Gy (three samples per dose) and calculated the CoV for each dosimeter concentration. Also, we used different masses of DNA strands (1, 2, 8, 16, 24, and 32 µg) to attach to the same volume of magnetic beads (100 µL) to explore how this affects the cost of the dosimeter. Results: The lowest CoV was produced for the highest concentration of dosimeter (100 µL of PBS), which created CoV of 2.0 and 1.0% for 25 and 50 Gy, respectively. We found that the lowest production cost for the dosimeter occurs by attaching 16 µg of DNA strands with 100 µL of beads. Conclusion: : We optimized the fabrication of the DNA dosimeter to produce low CoV and cost, but we still need to explore ways to further improve the dosimeter for use at lower doses. This work was supported in part by Yarmouk University (Irbid, Jordan) and CPRIT (RP140105)

  17. Investigations of interference between electromagnetic transponders and wireless MOSFET dosimeters: a phantom study.

    Science.gov (United States)

    Su, Zhong; Zhang, Lisha; Ramakrishnan, V; Hagan, Michael; Anscher, Mitchell

    2011-05-01

    To evaluate both the Calypso Systems' (Calypso Medical Technologies, Inc., Seattle, WA) localization accuracy in the presence of wireless metal-oxide-semiconductor field-effect transistor (MOSFET) dosimeters of dose verification system (DVS, Sicel Technologies, Inc., Morrisville, NC) and the dosimeters' reading accuracy in the presence of wireless electromagnetic transponders inside a phantom. A custom-made, solid-water phantom was fabricated with space for transponders and dosimeters. Two inserts were machined with positioning grooves precisely matching the dimensions of the transponders and dosimeters and were arranged in orthogonal and parallel orientations, respectively. To test the transponder localization accuracy with/without presence of dosimeters (hypothesis 1), multivariate analyses were performed on transponder-derived localization data with and without dosimeters at each preset distance to detect statistically significant localization differences between the control and test sets. To test dosimeter dose-reading accuracy with/without presence of transponders (hypothesis 2), an approach of alternating the transponder presence in seven identical fraction dose (100 cGy) deliveries and measurements was implemented. Two-way analysis of variance was performed to examine statistically significant dose-reading differences between the two groups and the different fractions. A relative-dose analysis method was also used to evaluate transponder impact on dose-reading accuracy after dose-fading effect was removed by a second-order polynomial fit. Multivariate analysis indicated that hypothesis 1 was false; there was a statistically significant difference between the localization data from the control and test sets. However, the upper and lower bounds of the 95% confidence intervals of the localized positional differences between the control and test sets were less than 0.1 mm, which was significantly smaller than the minimum clinical localization resolution of 0

  18. Automated personal dosimetry monitoring system for NPP

    International Nuclear Information System (INIS)

    Chanyshev, E.; Chechyotkin, N.; Kondratev, A.; Plyshevskaya, D.

    2006-01-01

    Full text: Radiation safety of personnel at nuclear power plants (NPP) is a priority aim. Degree of radiation exposure of personnel is defined by many factors: NPP design, operation of equipment, organizational management of radiation hazardous works and, certainly, safety culture of every employee. Automated Personal Dosimetry Monitoring System (A.P.D.M.S.) is applied at all nuclear power plants nowadays in Russia to eliminate the possibility of occupational radiation exposure beyond regulated level under different modes of NPP operation. A.P.D.M.S. provides individual radiation dose registration. In the paper the efforts of Design Bureau 'Promengineering' in construction of software and hardware complex of A.P.D.M.S. (S.H.W. A.P.D.M.S.) for NPP with PWR are presented. The developed complex is intended to automatize activities of radiation safety department when caring out individual dosimetry control. The complex covers all main processes concerning individual monitoring of external and internal radiation exposure as well as dose recording, management, and planning. S.H.W. A.P.D.M.S. is a multi-purpose system which software was designed on the modular approach. This approach presumes modification and extension of software using new components (modules) without changes in other components. Such structure makes the system flexible and allows modifying it in case of implementation a new radiation safety requirements and extending the scope of dosimetry monitoring. That gives the possibility to include with time new kinds of dosimetry control for Russian NPP in compliance with IAEA recommendations, for instance, control of the equivalent dose rate to the skin and the equivalent dose rate to the lens of the eye S.H.W. A.P.D.M.S. provides dosimetry control as follows: Current monitoring of external radiation exposure: - Gamma radiation dose measurement using radio-photoluminescent personal dosimeters. - Neutron radiation dose measurement using thermoluminescent

  19. Dose response of thin-film dosimeters irradiated with 80-120 keV electrons

    DEFF Research Database (Denmark)

    Helt-Hansen, J.; Miller, A.; Sharpe, P.

    2005-01-01

    Thin-film dosimeters (Riso B3 and alanine films) were irradiated at 10 MeV and 80-120 keV electron accelerators, and it has been shown that the radiation response of the dosimeter materials (the radiation chemical yields) are constant at these irradiation energies. However, dose gradients within ...... are present within the dosimeter. (C) 2005 Elsevier Ltd. All rights reserved....

  20. Biological monitoring of radiation exposure

    Science.gov (United States)

    Horneck, G.

    1998-11-01

    Complementary to physical dosimetry, biological dosimetry systems have been developed and applied which weight the different components of environmental radiation according to their biological efficacy. They generally give a record of the accumulated exposure of individuals with high sensitivity and specificity for the toxic agent under consideration. Basically three different types of biological detecting/monitoring systems are available: (i) intrinsic biological dosimeters that record the individual radiation exposure (humans, plants, animals) in measurable units. For monitoring ionizing radiation exposure, in situ biomarkers for genetic (e.g. chromosomal aberrations in human lymphocytes, germ line minisatellite mutation rates) or metabolic changes in serum, plasma and blood (e.g. serum lipids, lipoproteins, lipid peroxides, melatonin, antibody titer) have been used. (ii) Extrinsic biological dosimeters/indicators that record the accumulated dose in biological model systems. Their application includes long-term monitoring of changes in environmental UV radiation and its biological implications as well as dosimetry of personal UV exposure. (iii) Biological detectors/biosensors for genotoxic substances and agents such as bacterial assays (e.g. Ames test, SOS-type test) that are highly sensitive to genotoxins with high specificity. They may be applicable for different aspects in environmental monitoring including the International Space Station.

  1. Individual monitoring dosimetry in Europe

    International Nuclear Information System (INIS)

    Menzel, H.G.

    1991-01-01

    This report discusses the various types of individual monitoring systems presently in use within the European community and neutron dosimetry research being coordinated by the EURADOS working group. Research is currently being conducted on nuclear track dosimeters, primarily with CR-39 (TM), and TLD-albedo dosimeters. Studies are being conducted on the energy and angular response of each type of dosimeter. Because the response of dosimeters depends on the energy of the neutrons, it is necessary to have spectral information to accurately assess the dose. Neutron energy spectrum measurements are being performed in typical work place environments. Work is also progressing on development of calibration sources which will be representative of the neutron energy spectrum found in typical neutron exposure situations. This work utilizes 14 MeV neutrons incident on a uranium block with various other filters. Research is also continuing on neutron dosimetry using tissue equivalent proportional counters and microdosimetric techniques. The results of intercomparisons between several different instruments are discussed. In addition to personnel dosimetry, these systems are being used to record the dose to passengers and flight crews aboard commercial aircraft

  2. H-Area Seepage Basins groundwater monitoring report

    International Nuclear Information System (INIS)

    Thompson, C.Y.

    1992-06-01

    During first quarter 1992, tritium, nitrate, nonvolatile beta, total alpha-emitting radium (radium-224 and radium-226), gross alpha, antimony, mercury, lead, tetrachloroethylene, arsenic, and cadmium exceeded the US Environmental Protection Agency Primary Drinking Water Standards (PDWS) in groundwater samples from monitoring wells at the H-Area Seepage Basins (HASB) at the Savannah River Site. This report presents and discusses the groundwater monitoring results in the H-Area for first quarter 1992

  3. Liquid polymers for using in a holographic ionizing radiation dosimeter

    International Nuclear Information System (INIS)

    Nicolau-Rebigan, S.

    1979-01-01

    Some liquid polymeric systems for using in the holographic ionizing radiation dosimeter are presented. It is shown that the action of radiation on polymers leads to the destruction of the polymeric chains or to perform them, the both processes being applied in radiation dosimetry. Some advantages of the holographic dosimeter are outlined comparatively with those common used. (author)

  4. A polymeric dosimeter film based on optically-stimulated luminescence for dose measurements below 1 kGy

    International Nuclear Information System (INIS)

    Kovacs, A.; Baranyai, M.; Wojnarovits, L.; Slezsak, I.; McLaughlin, W.L.; Miller, S.D.; Miller, A.; Fuochi, P.G.; Lavalle, M.

    1999-01-01

    A new potential dosimetry system 'Sunna' containing a microcrystalline dispersion of an optically-stimulated fluor in a plastic matrix has been recently developed to measure and image high doses. Our previous investigations have revealed that the new dosimeter system is capable of measuring absorbed doses in the dose range of 1-100 kGy. The optically-stimulated luminescence (OSL) analysis is based on the blue light stimulation of the colour center states produced upon irradiation, and the intensity of the resulting red-light emission is used to measure absorbed dose. This analysis is carried out with a simple table-top fluorimeter developed for this purpose having also the ability to calculate the mathematical formula of the calibration function. The Sunna dosimeter was recently investigated for potential use in lower dose range below 1 kGy. These investigations have shown that the film is suitable for measuring doses in the range of 1-1000 Gy for both electron and gamma radiation. To test the applicability of the film, its reproducibility, stability, sensitivity to ambient and UV light and irradiation temperature were measured. The stability of the dosimeter was investigated by monitoring the change of the OSL signal with storage time after irradiation. Further experiments proved the homogeneity of the film with respect to thickness variation, and limited differences in its response were found between batches. (author)

  5. Design of Interrogation Protocols for Radiation Dose Measurements Using Optically-Stimulated Luminescent Dosimeters.

    Science.gov (United States)

    Abraham, Sara A; Kearfott, Kimberlee J; Jawad, Ali H; Boria, Andrew J; Buth, Tobias J; Dawson, Alexander S; Eng, Sheldon C; Frank, Samuel J; Green, Crystal A; Jacobs, Mitchell L; Liu, Kevin; Miklos, Joseph A; Nguyen, Hien; Rafique, Muhammad; Rucinski, Blake D; Smith, Travis; Tan, Yanliang

    2017-03-01

    Optically-stimulated luminescent dosimeters are capable of being interrogated multiple times post-irradiation. Each interrogation removes a fraction of the signal stored within the optically-stimulated luminescent dosimeter. This signal loss must be corrected to avoid systematic errors in estimating the average signal of a series of optically-stimulated luminescent dosimeter interrogations and requires a minimum number of consecutive readings to determine an average signal that is within a desired accuracy of the true signal with a desired statistical confidence. This paper establishes a technical basis for determining the required number of readings for a particular application of these dosimeters when using certain OSL dosimetry systems.

  6. An assessment of radiotherapy dosimeters based on CVD grown diamond

    International Nuclear Information System (INIS)

    Ramkumar, S.; Buttar, C.M.; Conway, J.; Whitehead, A.J.; Sussman, R.S.; Hill, G.; Walker, S.

    2001-01-01

    Diamond is potentially a very suitable material for use as a dosimeter for radiotherapy. Its radiation hardness, the near tissue equivalence and chemical inertness are some of the characteristics of diamond, which make it well suited for its application as a dosimeter. Recent advances in the synthesis of diamond by chemical vapour deposition (CVD) technology have resulted in the improvement in the quality of material and increased its suitability for radiotherapy applications. We report in this paper, the response of prototype dosimeters based on two different types (CVD1 and CVD2) of CVD diamond to X-rays. The diamond devices were assessed for sensitivity, dependence of response on dose and dose rate, and compared with a Scanditronix silicon photon diode and a PTW natural diamond dosimeter. The diamond devices of CVD1 type showed an initial increase in response with dose, which saturates after ∼6 Gy. The diamond devices of CVD2 type had a response at low fields ( 1162.8 V/cm), the CVD2-type devices showed polarisation and dose-rate dependence. The sensitivity of the CVD diamond devices varied between 82 and 1300 nC/Gy depending upon the sample type and the applied voltage. The sensitivity of CVD diamond devices was significantly higher than that of natural diamond and silicon dosimeters. The results suggest that CVD diamond devices can be fabricated for successful use in radiotherapy applications

  7. Sensitive color dosimeters using photochromic diarylethenes

    International Nuclear Information System (INIS)

    Irie, Setsuko; Irie, Masahiro

    2008-01-01

    Various types of color dosimeters are conveniently used for estimating absorbed dose in the radiation sterilization of biomedical materials. Diarylethenes with heterocyclic aryl groups are extensively studied for the applications to the optoelectronic devices, such as optical memory media and photowitching devices because of their thermally irreversible and fatigue-resistant properties. The colors of diarylethenes never fade in the dark conditions. The thermally stable dithienylethene derivatives are applied to sensitive color dosimeters. Upon γ-irradiation, polystyrene films containing diarylethene derivatives, such as 1,2-bis(2-methyl-5-phenyl-3-thienyl) perfluorocyclopentene 1 or 1,2-bis(2,5-dimethyl-3-thienyl) perfluorocyclopentene 2, and fluorescent metal complexes turned blue or red. Even if the absorbed dose was as small as 10 Gy, a clear color change was observed. (author)

  8. Effect of the temperature and relative humidity in dosemeters used for personnel monitoring

    International Nuclear Information System (INIS)

    Antonio Filho, J.

    1982-12-01

    The systematics of the combined effect of temperature and humidity on photographic dosimeters of the type Agfa-Gevaert, Kodak type II, III and the thermoluminescent dosimeters LiF:Mg,Ti (TLD-100, Harshaw), D-CaSO 4 :Dy-0,4 (Teledyne), e CaSO 4 :Dy+NaCl (IPEN), used in personal monitoring in Brazil was investigated, in the temperature range of 20 0 C to 50 0 C and relative humidity of 65% to 95%, in order to determine the best manner of utilization of these detectors in Brazilian climatic conditions. The dosimeters were studied in different forms of packing-sheet such as aluminezed paper and polyethylene. For the determination of the systematics, the dosimeters were irradiated in three conditions: before, during and after of storage in climatic chambers to a maximum period of 60 days. It was found that the dosimetric filmes and thermoluminescent dosimeter CaSO 4 :Dy+NaCl without protection, presented a high dependence to temperature and humidity, and when protected presented good results. Therefore, the best manner of utilization of these monitors in environments with relative humidity and temperature greater them 75% and 30 0 C respectively, is achieved with the protection of aluminized paper. The LiF:Mg,Ti and D+CaSO 4 :Dy-0,4 dosimeters can be utilized in their original form because they presented low dependence with humidity and temperature in the range studied. (Author) [pt

  9. Effect of pH grade on polymer-gel dosimeter and its brachytherapy application

    International Nuclear Information System (INIS)

    Spevacek, V.; Hrbacek, J.; Dvorak, P.; Cechak, T.; Novotny, J.

    2003-01-01

    To evaluate impact of pH grade on characteristics of polymer-gel dosimeter and its application in dose distribution verification in brachytherapy. A polymer-gel dosimeter based on radiation induced polymerization and crosslinking of acrylic monomers (acrylic acid, N,N' methylen-bis-acrylamide) was investigated with respect to its pH grade. pH grade of a dosimeter was varied by concentration of natrium hydroxide. Afterwards, dosimeter was split into several samples which were uniformly irradiated with Co-60 gamma rays. The range of doses applied was usually from 0 to 50 Gy with the main interest in region up to 20 Gy. Evaluation of dosimeter dose response was performed using MRI (T2). Dose response curves obtained were evaluated with respect to pH grade as a parameter. In parallel, there was studied temperature resistance (melting temperature) of gels with various pH grade. pH grade modified polymer-gel dosimeter was then used to compare dose distribution calculated with brachytherapy treatment planning system for simple irradiation geometry with Ir-192 HDR source. Additionaly, Monte Carlo calculated data were also included in the brachytherapy study. There was observed effect of pH grade on dose-response curve parameters (slope of linear fit, background response, linear range and maximum measurable dose). In general, the lower pH grade the higher sensitivity. Another positive effect of decreased pH grade is significantly higher maximum measurable dose. Maximum melting temperature of a gel was observed with pH grade between 3.5 and 4. For both higher and lower pH grades the melting temperature was lower. Using pH modified polymer-gel dosimeter simple brachytherapy dose distribution was measured and compared with calculated and Monte Carlo simulated data. There was observed strong dependence of dose-response relationship on pH grade of polymer-gel dosimeter resulting in significant improvement of dosimeter characteristics, namely sensitivity, applicable range of

  10. Characterization of MOSFET dosimeters for low-dose measurements in maxillofacial anthropomorphic phantoms.

    Science.gov (United States)

    Koivisto, Juha H; Wolff, Jan E; Kiljunen, Timo; Schulze, Dirk; Kortesniemi, Mika

    2015-07-08

    The aims of this study were to characterize reinforced metal-oxide-semiconductor field-effect transistor (MOSFET) dosimeters to assess the measurement uncertainty, single exposure low-dose limit with acceptable accuracy, and the number of exposures required to attain the corresponding limit of the thermoluminescent dosimeters (TLD). The second aim was to characterize MOSFET dosimeter sensitivities for two dental photon energy ranges, dose dependency, dose rate dependency, and accumulated dose dependency. A further aim was to compare the performance of MOSFETs with those of TLDs in an anthropomorphic phantom head using a dentomaxillofacial CBCT device. The uncertainty was assessed by exposing 20 MOSFETs and a Barracuda MPD reference dosimeter. The MOSFET dosimeter sensitivities were evaluated for two photon energy ranges (50-90 kVp) using a constant dose and polymethylmethacrylate backscatter material. MOSFET and TLD comparative point-dose measurements were performed on an anthropomorphic phantom that was exposed with a clinical CBCT protocol. The MOSFET single exposure low dose limit (25% uncertainty, k = 2) was 1.69 mGy. An averaging of eight MOSFET exposures was required to attain the corresponding TLD (0.3 mGy) low-dose limit. The sensitivity was 3.09 ± 0.13 mV/mGy independently of the photon energy used. The MOSFET dosimeters did not present dose or dose rate sensitivity but, however, presented a 1% decrease of sensitivity per 1000 mV for accumulated threshold voltages between 8300 mV and 17500 mV. The point doses in an anthropomorphic phantom ranged for MOSFETs between 0.24 mGy and 2.29 mGy and for TLDs between 0.25 and 2.09 mGy, respectively. The mean difference was -8%. The MOSFET dosimeters presented statistically insignificant energy dependency. By averaging multiple exposures, the MOSFET dosimeters can achieve a TLD-comparable low-dose limit and constitute a feasible method for diagnostic dosimetry using anthropomorphic phantoms. However, for single in

  11. Characterization of MOSFET dosimeters for low‐dose measurements in maxillofacial anthropomorphic phantoms

    Science.gov (United States)

    Wolff, Jan E.; Kiljunen, Timo; Schulze, Dirk; Kortesniemi, Mika

    2015-01-01

    The aims of this study were to characterize reinforced metal‐oxide‐semiconductor field‐effect transistor (MOSFET) dosimeters to assess the measurement uncertainty, single exposure low‐dose limit with acceptable accuracy, and the number of exposures required to attain the corresponding limit of the thermoluminescent dosimeters (TLD). The second aim was to characterize MOSFET dosimeter sensitivities for two dental photon energy ranges, dose dependency, dose rate dependency, and accumulated dose dependency. A further aim was to compare the performance of MOSFETs with those of TLDs in an anthropomorphic phantom head using a dentomaxillofacial CBCT device. The uncertainty was assessed by exposing 20 MOSFETs and a Barracuda MPD reference dosimeter. The MOSFET dosimeter sensitivities were evaluated for two photon energy ranges (50–90 kVp) using a constant dose and polymethylmethacrylate backscatter material. MOSFET and TLD comparative point‐dose measurements were performed on an anthropomorphic phantom that was exposed with a clinical CBCT protocol. The MOSFET single exposure low dose limit (25% uncertainty, k=2) was 1.69 mGy. An averaging of eight MOSFET exposures was required to attain the corresponding TLD (0.3 mGy) low‐dose limit. The sensitivity was 3.09±0.13 mV/mGy independently of the photon energy used. The MOSFET dosimeters did not present dose or dose rate sensitivity but, however, presented a 1% decrease of sensitivity per 1000 mV for accumulated threshold voltages between 8300 mV and 17500 mV. The point doses in an anthropomorphic phantom ranged for MOSFETs between 0.24 mGy and 2.29 mGy and for TLDs between 0.25 and 2.09 mGy, respectively. The mean difference was −8%. The MOSFET dosimeters presented statistically insignificant energy dependency. By averaging multiple exposures, the MOSFET dosimeters can achieve a TLD‐comparable low‐dose limit and constitute a feasible method for diagnostic dosimetry using anthropomorphic phantoms. However

  12. Photon energy readings in OSL dosimeter filters: an application to retrospective dose estimation for nuclear medicine workers.

    Science.gov (United States)

    Villoing, Daphnée; Kitahara, Cari M; Passmore, Christopher; Simon, Steven L; Yoder, R Craig

    2018-06-19

    This work investigates the applicability of using data from personal monitoring dosimeters to assess photon energies to which medical workers were exposed. Such determinations would be important for retrospective assessments of organ doses to be used in occupational radiation epidemiology studies, particularly in the absence of work history or other information regarding the energy of the radiation source. Monthly personal dose equivalents and filter ratios under two different metallic filters contained in the Luxel+® dosimeter were collected from Landauer, Inc. from 19 nuclear medicine (NM) technologists employed by three medical institutions, the institution A only performing traditional NM imaging (primarily using 99mTc) and institutions B and C also performing positron emission tomography (PET, using 18F). Calibration data of the Luxel+® dosimeter for various X-ray spectra were used to establish ranges of filter ratios from 1.1 to 1.6 for 99mTc and below 1.1 for 18F. Median filter ratios were 1.33 (Interquartile range (IQR), 0.15) for institution A, 1.08 (IQR, 0.16) for institution B, and 1.08 (IQR, 0.14) for institution C. The distributions of these filter ratios were statistically-significantly different between the institution A only performing traditional NM imaging and institutions B and C also performing PET imaging. In this proof-of-concept study, filter ratios from personal dosimeters were used to assess differences in photon energies to which NM technologists were exposed. Dosimeters from technologists only performing traditional NM procedures mostly showed Al/Cu filter ratios above 1.2, those likely performing only PET in a particular month had filter ratios below 1.1, and those which showed filter ratios between 1.1 and 1.2 likely came from technologists rotating between traditional NM and PET imaging in the same month. These results suggest that it is possible to distinguish technologists who

  13. Response of a neutron monitor area with TLDs pairs

    Energy Technology Data Exchange (ETDEWEB)

    Guzman G, K. A.; Borja H, C. G.; Valero L, C.; Hernandez D, V. M.; Vega C, H. R. [Universidad Autonoma de Zacatecas, Unidad Academica de Estudios Nucleares, Calle Cipres No. 10, Fracc. La Penuela, 98068 Zacatecas (Mexico); Gallego, E.; Lorente, A., E-mail: ing_karen_guzman@yahoo.com.mx [Universidad Politecnica de Madrid, Departamento de Ingenieria Nuclear, Jose Gutierrez Abascal 2, E-28006 Madrid (Spain)

    2011-10-15

    The response of a passive neutron monitor area has been calculated using the Monte Carlo code MCNP5. The response was the amount of n({sup 6}Li, T){alpha} reactions occurring in a TLD-600 located at the center of a cylindrical polyethylene moderator. Fluence, (n, a) and H*(10) responses were calculated for 47 monoenergetic neutron sources. The H*(10) relative response was compared with responses of commercially available neutron monitors being alike. Due to {sup 6}Li cross section (n, {alpha}) reactions are mainly produced by thermal neutrons, however TLD-600 is sensitive to gamma-rays; to eliminate the signal due to photons monitor area was built to hold 2 pairs of TLD-600 and 2 pairs of TLD-700, thus from the difference between TLD-600 and TLD-700 readouts the net signal due to neutrons is obtained. The monitor area was calibrated at the Universidad Politecnica de Madrid using a {sup 241}AmBe neutron source; net TLD readout was compared with the H*(10) measured with a Bert hold Lb-6411. Performance of the neutron monitor area was determined through two independent experiments, in both cases the H*(10) was statistically equal to H*(10) measured with a Bert hold Lb-6411. Neutron monitor area with TLDs pairs can be used in working areas with intense, mixed and pulsed radiation fields. (Author)

  14. Studies on reduction of dosimeter used in the product dose mapping process at Sinagama Plant

    International Nuclear Information System (INIS)

    Sofian Ibrahim; Syuhada Ramli; Cosmos George; Zarina Mohd Nor; Kamarudin Buyong; Shahidan Yob; Nor Ishadi Ismail; Mohd Sidek Othman; Ahsanulkhaliqin Abdul Wahab; Mohd Khairul Azfar Ramli

    2012-01-01

    Product dose mapping is the determination of the best product loading configuration which will be used during routine sterilization. In product dose mapping, dosimeters are placed throughout products at strategic locations to determine the zones of minimum and maximum dose. On previous Sinagama's product dose mapping method, a total of 240 unit's ceric-cerous dosimeter been used for a tote. Based on the data obtained from Irradiator Dose Mapping Report in 2004 and data from recent studies, the number of dosimeter to be used in product dose mapping can be reduced to 28 units without sacrificing precision and accuracy of the dose mapping results. This also led changes of the placing dosimeter method from Plane system to Coordinate system. Reduction of 88 % on dosimeters usage will directly reduce the cost of expenses on dosimeter, time and labor. (author)

  15. Angular dependence of dose equivalent response of an albedo neutron dosimeter

    International Nuclear Information System (INIS)

    Torres, B.A.; Boswell, E.; Schwartz, R.B.

    1994-01-01

    The ANSI provides procedures for testing the performance of dosimetry services. Although neutron dose equivalent angular response studies are not now mandated, future standards may well require that such studies be performed. Current studies with an albedo dosimeter will yield information regarding the angular dependence of dose equivalent response for this type of personnel dosimeter. Preliminary data for bare 252 Cf fluences show a marked decrease in dosimeter reading with increasing angle. The response decreased by an approximate factor of four. For the horizontal orientation, the same response was noted from both positive and negative angles. However, for the vertical orientation, the response was unexplainably assymetric. We are also examining the response of the personnel badge in moderated 252 Cf fluences. Responses from the moderated and unmoderated 252 Cf fields and theoretical calculations of the neutron angular response will be compared. This information will assist in building a data base for future comparisons of neutron angular responses with other neutron albedo dosimeters and phantoms

  16. Development of real time personal neutron dosimeter with two silicon detectors

    Energy Technology Data Exchange (ETDEWEB)

    Nakamura, T.; Tsujimura, N. [Tohoku Univ., Cyclotron and Radioisotope Center, Aoba, Aramaki, Aoba-ku (Japan); Yamano, T. [Tokyo Factory, Fuji Electric Co. Ltd., Tokyo (Japan)

    1992-07-01

    We developed a real time personal neutron dosimeter by using two types of silicon p-n junction detectors, thermal neutron sensor and fast neutron sensor. The thermal neutron sensor which is {sup 10}B doped n-type silicon with a polyethylene radiator mainly counts neutrons of energy front thermal to I MeV, and the fast neutron sensor which is p-type silicon with a polyethylene radiator is sensitive to neutrons above I MeV. The neutron sensitivity measurements revealed that the dosimeter has a rather flat response for dose equivalent from thermal to 15 MeV, excluding a drop from 50 keV to I MeV. In order to get conversion factor from counts to dose equivalent as accurately as possible, we performed the field test of the dosimeter calibration in several neutron-generating fields. By introducing the two-group dose estimation method, this dosimeter can give the neutron dose equivalent within about 50% errors. (author)

  17. PRESAGE® as a solid 3-D radiation dosimeter: A review article

    Science.gov (United States)

    Khezerloo, Davood; Nedaie, Hassan Ali; Takavar, Abbas; Zirak, Alireza; Farhood, Bagher; Movahedinejhad, Hadi; Banaee, Nooshin; Ahmadalidokht, Isa; Knuap, Courtney

    2017-12-01

    Radiation oncology has been rapidly improved by the application of new equipment and techniques. With the advent of new complex and precise radiotherapy techniques such as intensity modulated radiotherapy, stereotactic radiosurgery, and volumetric modulated arc therapy, the demand for an accurate and feasible three-dimensional (3-D) dosimetry system has increased. The most important features of a 3-D dosimeter, apart from being precise, accurate and reproducible, include also its low cost, feasibility, and availability. In 2004 a new generation of solid plastic dosimeters which demonstrate a radiochromic response to ionizing radiation was introduced. PRESAGE® plastic dosimeter lacks the limitations of previous Ferric and polymer plastic 3-D dosimeters such as diffusion, sensitivity to oxygen, fabrication problems, scanning and read out challenges. In this decade, a large number of efforts have been carried out to enhance PRESAGE® structure and scanning methods. This article attempts to review and reflect on the results of these investigations.

  18. Solar UV exposure among outdoor workers in Denmark measured with personal UV-B dosimeters

    DEFF Research Database (Denmark)

    Grandahl, Kasper; Mortensen, Ole Steen; Sherman, David Zim

    2017-01-01

    radiation exposure are needed to help resolve this problem. This can be done using personal ultraviolet radiation dosimeters. Methods: We consider technical and practical feasibility of measuring individual solar ultraviolet exposure at work and leisure in professions with different á priori temporal high......-level outdoor worktime, using aluminium gallium nitride (AlGaN) photodiode detector based personal UV-B dosimeters. Essential technical specifications including the spectral and angular responsivity of the dosimeters are described and pre-campaign dosimeter calibration applicability is verified. The scale...... with our specialist knowledge as occupational physicians. Conclusions: Large-scale use of personal UV-B dosimeters for measurement of solar ultraviolet radiation exposure at work and leisure in Denmark is indeed feasible from a technical and practical viewpoint. Samples of exposure data shown support...

  19. Response of TLD-100 LiF dosimeters for X-rays of low energies

    International Nuclear Information System (INIS)

    Bonzi, E. V.; Mainardi, R. T.

    2011-10-01

    In diverse practical applications as the existent in radiological clinics, industrial facilities and research laboratories, the solid state dosimeters are used for the measure of the different types of ionizing radiations. At the present time dosimeters are manufactured with different types of materials that present thermoluminescent properties, to the effects of determining the absorbed radiation dose. Under these conditions, the radiation dose is determined integrated in all the range of energies of the beam of X-rays, since it assumes that the response of these dosimeters is lineal with the energy of the photons or radiant particles. Because interest exists in advancing in the development of a determination method in the way of the X-rays spectrum emitted by a tube of those used in diagnostic or therapy, we have measured the response of TLD-100 LiF dosimeters for low energies, minor at 60 keV, for a several group of these dosimeters. (Author)

  20. NRC TLD Direct Radiation Monitoring Network

    International Nuclear Information System (INIS)

    Struckmeyer, R.

    1994-03-01

    This report presents the results of the NRC Direct Radiation Monitoring Network for the fourth quarter of 1993. It provides the ambient radiation levels measured in the vicinity of 75 sites throughout the United States. In addition, it describes the equipment used, monitoring station selection criteria, characterization of the dosimeter response, calibration procedures, statistical methods, intercomparison, and quality assurance program

  1. NRC TLD Direct Radiation Monitoring Network

    International Nuclear Information System (INIS)

    Struckmeyer, R.; McNamara, N.

    1993-03-01

    This report present the results of the NRC Direct Radiation Monitoring Network for the fourth quarter of 1992. It provides the ambient radiation levels measured in the vicinity of 75 sites throughout the United States. In addition, it describes the equipment used, monitoring station selection criteria, characterization of the dosimeter response, calibration procedures, statistical methods, intercomparison, and quality assurance program

  2. Electron-energy deposition in skin and thermoluminescence dosimeters

    International Nuclear Information System (INIS)

    Mei, G.T.Y.

    1986-01-01

    The primary object of this study was to investigate the relations between dosimeter response and skin dose resulting from beta-particle irradiation. This object was achieved by combining evaluation of beta-source energy spectra, calculation of flux energy spectra, and employment of a Monte-Carlo electron-transport computer program for determination of depth-dose distribution in multislab geometries. Intermediate results from three steps of evaluation were compared individually with experimental data or with other theoretical results and showed excellent agreement. The combined method is applicable for the electron agreement. The combined method is applicable for the electron energy range of 1 keV to 5 MeV for both monoenergetic electrons and energy-distributed electrons. Determination of dosimeter response - skin dose relationships for homogeneous atmospheric beta-particle sources and for two specific configurations of LiF TLD's have been carried out in this study. Information based on these calculations is of value in designing beta-particle dosimeters as well as in assessing potential occupational and public health risks associated with the nuclear power industry

  3. The NRPB's new dosimeter and dose record keeping services

    International Nuclear Information System (INIS)

    Dennis, J.A.; Marshall, T.O.; Shaw, K.B.

    1976-01-01

    A new automated dosimeter and record keeping service which the National Radiological Protection Board (UK) intends to introduce in 1977 is described. The automated system, based on a thermoluminescent dosimeter, will be linked to a fully computerised record keeping system with automatic printing of dose records and Transfer Records operated at its Headquarters at Harwell. The new system will dispense with much manual labour which in the past has introduced inevitable errors and incurred increasing costs. (U.K.)

  4. Valine-spectrophotometric readout dosimeter (1 Gy-50 kGy)

    International Nuclear Information System (INIS)

    Nilekani, S.R.; Gupta, B.L.

    2005-01-01

    In this method 20 mg unirradiated/irradiated L-valine powder [(CH 3 ) 2 CH.NH 2 CH.COOH] is dissolved in 10 ml of a solution containing 4x10 -4 mol dm -3 Fe 2+ and 2.5x10 -4 mol dm -3 xylenol orange (XO) in aerated aqueous 0.060 mol dm -3 sulphuric acid (FX). The plot of absorbance at 550 nm against dose is non linear. A dose of 1-50 kGy can be measured. However, dosimeter can be sensitized in the dose range of 1 to 16 kGy by dissolving 50-mg valine powder in 10 ml of a solution which contains 5x10 -4 mol dm -3 Fe 2+ and 3x10 -4 mol dm -3 XO in aerated aqueous 0.065 mol dm -3 sulphuric acid. The plot of absorbance at 549 nm against dose is non-linear. However, dosimeter shows linear response when 500 mg unirradiated/irradiated L-valine powder is dissolved in 10 ml of a solution containing 7.5x10 -4 mol dm -3 Fe 2+ and 3x10 -4 mol dm -3 XO in aerated aqueous 0.25 mol dm -3 sulphuric acid. The plot of absorbance at 557 nm against dose is linear in the dose range of 20-400Gy and doses down to about 1 Gy can be measured using 10-cm path cells. Response of the dosimeter is independent of irradiation temperature in the temperature range 20-50 deg C. Irradiated valine powder is stable for about 1 month. The reproducibility of the method is better than ±2%. This dosimeter is very useful as transfer dosimeter for food irradiation and radiation sterilization

  5. Ionizing radiation M.O.S. dosimeters: sensibility and stability

    International Nuclear Information System (INIS)

    Gessinn, F.

    1993-12-01

    This thesis is a contribution to the study of the ionizing radiation responsivity of P.O.M.S. dosimeters. Unlike the development of processing hardening techniques, our works goal were to increase, on the one hand, the M.O.S. dosimeters sensitivity in order to detect small radiation doses and on the other hand, the stability with time and temperature of the devices, to minimize the absorbed-dose estimation errors. With this aim in mind, an analysis of all processing parameters has been carried out: the M.O.S. dosimeter sensitivity is primarily controlled by the gate oxide thickness and the irradiation electric field. Thus, P.M.O.S. transistors with 1 and 2 μm thick silica layers have been fabricated for our experiments. The radiation response of our devices in the high-field mode satisfactorily fits a D ox 2 power law. The maximum sensitivity achieved (9,2 V/Gy for 2μm devices) is close to the ideal value obtained when considering only an unitary carrier-trapping level, and allows to measure about 10 -2 Gy radiation doses. Read-time stability has been evaluated under bias-temperature stress conditions: experiments underscore slow fading, corresponding to 10 -3 Gy/h. The temperature response has also been studied: the analytical model we have developed predicts M.O.S. transistors threshold voltage variations over the military specifications range [-50 deg. C, + 150 deg. C]. Finally, we have investigated the possibilities of irradiated dosimeters thermal annealing for reusing. It appears clearly that radiation-induced damage annealing is strongly gate bias dependent. Furthermore, dosimeters radiation sensitivity seems not to be affected by successive annealings. (author). 146 refs., 58 figs., 9 tabs

  6. NRC TLD Direct Radiation Monitoring Network

    International Nuclear Information System (INIS)

    Struckmeyer, R.; McNamara, N.

    1991-12-01

    This report provides the status and results of the NRC Thermoluminescent Dosimeter (TLD) Direct Radiation Monitoring Network. It presents the radiation levels measured in the vicinity of NRC licensed facilities throughout the country for the third quarter of 1991

  7. Effect of the shape and size of dosimeters on the response of solid state/EPR dosimetry

    International Nuclear Information System (INIS)

    Yordanov, Nicola D.; Fabisiak, Slawomir; Lagunov, Oleg

    2006-01-01

    The influence of the shape and size of dosimeters used in solid state-EPR (SS/EPR) dosimetry on their response is reported. It is shown that for commonly used cylindrical (rod) shaped dosimeters of equal height, prepared of low (ε=<3) dielectric constant materials, linearity between their volume and the EPR response is observed when their diameter varies between 3 and 5mm. Further increase of the dosimeter's diameter is not recommended since the increased penetration of the dosimeter material into the electric component of the microwave field in the EPR cavity increases the dielectric losses and decreases the EPR response. In an attempt to improve the sensitivity of the SS/EPR dosimetry we have prepared and tested new, flat-shaped, dosimeters of low (ε∼2) dielectric constant materials which were found to exhibit: (i) linear EPR response within 1-5mm thickness; (ii) higher sensitivity than cylindrical dosimeters at equal sample volume; (iii) increased by ca. 270% EPR sensitivity at 5mm thickness compared to the cylindrical dosimeters with the same diameter (ca. 1.7 times increased sample volume). Using flat shape dosimeters of suitable size provides 2.7 times higher EPR sensitivity of single estimation

  8. Development study of a quality control for clinical dosimeters of radiotherapy

    International Nuclear Information System (INIS)

    Damatto, Willian B.; Potiens, Maria P.A.; Santos, Gelson P.; Vivolo, Vitor

    2011-01-01

    This paper presents the partial results of a scientific initiation which the main objective is the enhancement of the quality system of the dosimeter calibration laboratory (LCI-IPEN) on 60 Co gamma radiation to the International Atomic Energy Agency (IAEA), being this the new protocol denominated Calibration of Reference Dosimeters for External Beam Radiotherapy (Technical Reports Series 469). This paper is an actualization of the protocol Absorbed Dose Determination for External Beam Radiotherapy (Technical Reports Series 398). Therefore, in this paper it will presented the study on the clinical dosimeters composed of electrometer, wires, triaxial connectors, and thimble type ionization chamber - 0.60 cm 3 . (author)

  9. Radiation Dose Measurement Using Chemical Dosimeters

    International Nuclear Information System (INIS)

    Lee, Min Sun; Kim, Eun Hee; Kim, Yu Ri; Han, Bum Soo

    2010-01-01

    The radiation dose can be estimated in various ways. Dose estimates can be obtained by either experiment or theoretical analysis. In experiments, radiation impact is assessed by measuring any change caused by energy deposition to the exposed matter, in terms of energy state (physical change), chemical production (chemical change) or biological abnormality (biological change). The chemical dosimetry is based on the implication that the energy deposited to the matter can be inferred from the consequential change in chemical production. The chemical dosimetry usually works on the sample that is an aqueous solution, a biological matter, or an organic substance. In this study, we estimated absorbed doses by quantitating chemical changes in matter caused by radiation exposure. Two different chemical dosimeters, Fricke and ECB (Ethanol-Chlorobenzene) dosimeter, were compared in several features including efficacy as dose indicator and effective dose range

  10. Nevada Test 1999 Waste Management Monitoring Report, Area 3 and Area 5 radioactive waste management sites

    International Nuclear Information System (INIS)

    Yvonne Townsend

    2000-01-01

    Environmental monitoring data were collected at and around the Area 3 and Area 5 Radioactive Waste Management Sites (RWMSs) at the Nevada Test Site (NTS). These monitoring data include radiation exposure, air, groundwater, meteorology, vadose zone, and biota data. Although some of these media (radiation exposure, air, and groundwater) are reported in detail in other Bechtel Nevada reports (Annual Site Environmental Report [ASER], the National Emissions Standard for Hazardous Air Pollutants [NESHAP] report, and the Annual Groundwater Monitoring Report), they are also summarized in this report to provide an overall evaluation of RWMS performance and environmental compliance. Direct radiation monitoring data indicate that exposure at and around the RWMSs is not above background levels. Air monitoring data indicate that tritium concentrations are slightly above background levels, whereas radon concentrations are not above background levels. Groundwater monitoring data indicate that the groundwater in the alluvial aquifer beneath the Area 5 RWMS has not been affected by the facility. Meteorology data indicate that 1999 was a dry year: rainfall totaled 3.9 inches at the Area 3 RWMS (61 percent of average) and 3.8 inches at the Area 5 RWMS (75 percent of average). Vadose zone monitoring data indicate that 1999 rainfall infiltrated less than one foot before being returned to the atmosphere by evaporation. Soil-gas tritium data indicate very slow migration, and tritium concentrations in biota were insignificant. All 1999 monitoring data indicate that the Area 3 and Area 5 RWMSs are performing as expected at isolating buried waste

  11. Moisture Monitoring at Area G, Technical Area 54, Los Alamos National Laboratory, 2016 Status Report

    Energy Technology Data Exchange (ETDEWEB)

    Levitt, Daniel Glenn [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Birdsell, Kay Hanson [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Jennings, Terry L. [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); French, Sean B. [Los Alamos National Lab. (LANL), Los Alamos, NM (United States)

    2017-01-17

    Hydrological characterization and moisture monitoring activities provide data required for evaluating the transport of subsurface contaminants in the unsaturated and saturated zones beneath Area G, and for the Area G Performance Assessment and Composite Analysis. These activities have been ongoing at Area G, Technical Area 54 of the Los Alamos National Laboratory since waste disposal operations began in 1957. This report summarizes the hydrological characterization and moisture monitoring activities conducted at Area G. It includes moisture monitoring data collected from 1986 through 2016 from numerous boreholes and access tubes with neutron moisture meters, as well as data collected by automated dataloggers for water content measurement sensors installed in a waste disposal pit cover, and buried beneath the floor of a waste disposal pit. This report is an update of a nearly identical report by Levitt et al., (2015) that summarized data collected through early 2015; this report includes additional moisture monitoring data collected at Pit 31 and the Pit 38 extension through December, 2016. It also includes information from the Jennings and French (2009) moisture monitoring report and includes all data from Jennings and French (2009) and the Draft 2010 Addendum moisture monitoring report (Jennings and French, 2010). For the 2015 version of this report, all neutron logging data, including neutron probe calibrations, were investigated for quality and pedigree. Some data were recalculated using more defensible calibration data. Therefore, some water content profiles are different from those in the Jennings and French (2009) report. All of that information is repeated in this report for completeness. Monitoring and characterization data generally indicate that some areas of the Area G vadose zone are consistent with undisturbed conditions, with water contents of less than five percent by volume in the top two layers of the Bandelier tuff at Area G. These data also

  12. Fabrication and Optimization of a PAGATA Gel Dosimeter: Increasing the Melting Point of the PAGAT Gel Dosimeter with Agarose Additive

    Directory of Open Access Journals (Sweden)

    Bakhtiar Azadbakht

    2010-12-01

    Full Text Available Introduction: The PAGAT polymer gel dosimeter melts at 30 ˚C and even at room temperature during the summer, so it needs to be kept in a cool place such as a refrigerator. To increase the stability of the PAGAT gel, different amounts of agarose were added to the PAGAT gel composition and the PAGATA gel was manufactured. Material and Methods: The PAGATA gel vials were irradiated using a Co-60 machine. Then, the samples were evaluated using a 1.5 T Siemens MRI scanner. The ingredients of the PAGATA normoxic gel dosimeter were 4.5% N-N' methylen-bis-acrylamide, 4.5% acrylamide, 4.5% gelatine, 5 mM tetrakis (THPC, 0.01 mM hydroquinone (HQ, 0.5% agarose and 86% de-ionized water (HPLC. Results: Melting point and sensitivity of the PAGAT gel dosimeter with addition of 0.0, 0.3, 0.5, 1.0, 1.5 and 2.0% of agarose were measured, in which the melting points were increased to 30, 82, 86, 88, 89 and 90°C and their sensitivities found to be 0.113, 0.1059, 0.125, 0.122, 0.115 and 0.2  respectively. Discussion and Conclusions: Adding agarose increased the sensitivity and background R2 of the evaluated samples. The optimum amount of agarose was found to be 0.5% regarding these parameters and also the melting point of the gel dosimeter. A value of 0.5% agarose was found to be an optimum value considering the increase of sensitivity to 0.125 and melting point to 86°C but at the expense of increasing the background R2 to 4.530.

  13. Personnel radiation monitoring by thermoluminescence dosimetry (1995-96)

    International Nuclear Information System (INIS)

    Daw Mi Cho Cho; Daw Yi Yi Khin; Daw San San; U Maung Maung Tin; Daw Hla Hla Win

    2001-01-01

    Personnel radiation monitoring which is the dose assessment of individual doses from external radiation received by radiation workers has been carried out by Thermoluminescence Dosimetry system consisting of a Vinten Toledo TLD reader, LiF dosimeters and associated equipment. The exposed TLD dosimeters were measured by TLD reader and the dose evaluation and dose registration were done on personal computer. Due to the records of 1995-96, most of the radiation workers complied with the permissible dose recommended by IAEA and ICRP 60. (author)

  14. 2016 Waste Management Monitoring Report Area 3 and Area 5 Radioactive Waste Management Sites

    Energy Technology Data Exchange (ETDEWEB)

    Black, David [National Security Technologies, LLC. (NSTec), Mercury, NV (United States)

    2017-08-30

    Environmental monitoring data are collected at and around the Area 3 and Area 5 Radioactive Waste Management Sites (RWMSs) within the Nevada National Security Site (NNSS). These data include direct radiation exposure, as well as radiation from the air, groundwater, meteorology, and vadose zone. This report summarizes the 2016 environmental data to provide an overall evaluation of RWMS performance and to support environmental compliance and performance assessment (PA) activities. Some of these data (e.g., radiation exposure, air, and groundwater) are presented in other reports, developed by National Security Technologies, LLC Direct radiation monitoring data indicate exposure levels at the RWMSs are within the range of background levels measured at the NNSS. Slightly elevated exposure levels outside the Area 3 RWMS are attributed to nearby historical aboveground nuclear weapons tests. Air monitoring data show that tritium concentrations in water vapor and americium and plutonium concentrations in air particles are below Derived Concentration Standards for these radionuclides. Groundwater monitoring data indicate the groundwater in the uppermost aquifer beneath the Area 5 RWMS is not impacted by RWMS operations. Results of groundwater analysis from wells around the Area 5 RWMS were all below established investigation levels. Leachate samples collected from the leachate collection system at the mixed low-level waste cell were below established contaminant regulatory limits. During 2016, precipitation at the Area 3 RWMS was 8% below average, and precipitation at the Area 5 RWMS was 8% above average. Water balance measurements indicate that evapotranspiration from the vegetated weighing lysimeter dries the soil and prevents downward percolation of precipitation more effectively than evaporation as measured from the bare-soil weighing lysimeter. Vadose zone monitoring on Area 5 and Area 3 RWMS cell covers shows no evidence of precipitation percolating through the covers

  15. Nevada Test Site 2001 Waste Management Monitoring Report Area 3 and Area 5 Radioactive Waste Management Sites

    International Nuclear Information System (INIS)

    Y. E. Townsend

    2002-06-01

    Environmental monitoring data, subsidence monitoring data, and meteorology monitoring data were collected at and around the Area 3 and Area 5 Radioactive Waste Management Sites (RWMSs) at the Nevada Test Site (NTS) (refer to Figure 1). These monitoring data include radiation exposure, air, groundwater, meteorology, vadose zone, subsidence, and biota data. Although some of these media (radiation exposure, air, and groundwater) are reported in detail in other Bechtel Nevada (BN) reports (Annual Site Environmental Report [ASER], the National Emissions Standard for Hazardous Air Pollutants [NESHAP] report, and the Annual Groundwater Monitoring Report), they are also summarized in this report to provide an overall evaluation of RWMS performance and environmental compliance. Direct radiation monitoring data indicate that exposure at and around the RWMSs is not above background levels. Air monitoring data indicate that tritium concentrations are slightly above background levels. Groundwater monitoring data indicate that the groundwater in the uppermost aquifer beneath the Area 5 RWMS has not been affected by the facility. Meteorology data indicate that 2001 was an average rainfall year: rainfall totaled 150 mm (5.9 in) at the Area 3 RWMS and 120 mm (4.7 in) at the Area 5 RWMS. Vadose zone monitoring data indicate that 2001 rainfall infiltrated less than one meter (3 ft) before being returned to the atmosphere by evaporation. Soil-gas tritium monitoring data indicate slow subsurface migration, and tritium concentrations in biota were lower than in previous years. All 2001 monitoring data indicate that the Area 3 and Area 5 RWMSs are performing within expectations of the model and parameter assumptions for the facility performance assessments

  16. TH-C-19A-05: Evaluation of a New Reusable 3D Dosimeter

    International Nuclear Information System (INIS)

    Juang, T; Adamovics, J; Oldham, M

    2014-01-01

    Purpose: PRESAGE is a radiochromic plastic which has demonstrated strong potential for high resolution single-use 3D dosimetry. This study evaluates a new PRESAGE formulation (Presage-RU) in which the radiochromic response is reversible (the dosimeter optically clears after irradiation), enabling the potential for reusability. Methods: Presage-RU dose response and optical-clearing rates were evaluated in both small volume dosimeters (1×1×4.5cm) and a larger cylindrical dosimeter (8cm diameter, 4.5cm length). All dosimeters were allowed to fully optically clear in dark, room temperature conditions between irradiations. Dose response was determined by irradiating small volume samples from 0–8.0Gy and measuring change in optical density. The cylindrical dosimeter was irradiated with a simple 4-field box plan (parallel opposed pairs of 4cm×4cm AP-PA beams and 2cm×4cm lateral beams) to 20Gy. High resolution 3D dosimetry was achieved utilizing optical-CT readout. Readings were tracked up to 14 days to characterize optical clearing. Results: Initial irradiation yielded a response of 0.0119△OD/(Gy*cm) while two subsequent reirradiations yielded a lower but consistent response of 0.0087△OD/(Gy*cm). Strong linearity of dose response was observed for all irradiations. In the large cylindrical dosimeter, the integral dose within the high dose region exhibited an exponential decay in signal over time (halflife= 23.9 hours), with the dosimeter effectively cleared (0.04% of the initial signal) after 10 days. Subsequent irradiation resulted in 19.5% lower initial signal but demonstrated that the exponential clearing rate remained consistent. Results of additional subsequent irradiations will also be presented. Conclusion: This work introduces a new re-usable radiochromic dosimeter (Presage-RU) compatible with high resolution (sub-millimeter) 3D dosimetry. Sensitivity of the initial radiation was observed to be slightly higher than subsequent irradiations, but the

  17. Calibration of an ALBEDO termoluminiscent dosimeter for its use in personal dosimetry

    International Nuclear Information System (INIS)

    Diaz Bernal, E.; Molina Perez, D.; Cornejo Diaz, N.; Carrazana Gonzalez, J.

    1996-01-01

    The dosimetric studies began after the Radiological Individuals Surveillance Department from the Radiation Protection and Hygiene Center acquired the albedo thermoluminescent dosimeters model JR1104. This paper reviews the response of those dosimeters to the different spectrums and incidence angles of neutronic radiation

  18. SU-E-T-274: Does Atmospheric Oxygen Affect the PRESAGE Dosimeter?

    Energy Technology Data Exchange (ETDEWEB)

    Alqathami, M; Ibbott, G [UT MD Anderson Cancer Center, Houston, TX (United States); Blencowe, A [The University of South Australia, South Australia, SA (Australia)

    2015-06-15

    Purpose: To experimentally determine the influence of atmospheric oxygen on the efficiency of the PRESAGE dosimeter and its reporting system. Methods: Batches of the reporting system – a mixture of chloroform and leuchomalachite green dye – and PRESAGE were prepared in aerobic and anaerobic conditions. For anaerobic batches, samples were deoxygenated by bubbling nitrogen through the dosimeter precursors or reporting system for 10 min. The dosimeters and reporting systems were prepared in spectrophotometric cuvettes and glass vials, respectively, and were irradiated with 6 MV photons to various radiation doses using a clinical linear accelerator. Changes in optical density of the dosimeters and reporting system before and after irradiation were measured using a spectrophotometer. In addition, the concentrations of dissolved oxygen were measured using a dissolved oxygen meter. Results: The experiments revealed that oxygen has little influence on the characteristics of PRESAGE, with the radical initiator oxidizing the leucomalachite green even in the presence of oxygen. However, deoxygenation of the reporting system leads to an increase in sensitivity to radiation dose by ∼ 30% when compared to the non-deoxygenated system. A slight improvement in sensitivity (∼ 5%) was also achieved by deoxygenating the PRESAGE precursor prior to casting. Measurement of the dissolved oxygen revealed low levels (0.4 ppm) in the polyurethane precursor used to fabricate the dosimeters, as compared to water (8.6 ppm). In addition, deoxygenation had no effect on the retention of the post-response absorption value of the PRESAGE dosimeter. Conclusion: The results suggest that the presence of oxygen does not inhibit the radiochromic properties of the PRESAGE system. In addition, there were no observed changes in the dose linearity, absorption spectrum and post-response photofading characteristics of the PRESAGE under the conditions investigated.

  19. Study and characterization of dosimeter LiF:Mg,Cu,P for using in aeronautical dosimetry

    International Nuclear Information System (INIS)

    Flavia, Hanna; Federico, Claudio; Lelis, Odair; Pereira, Heloisa; Pereira, Marlon

    2014-01-01

    The effects of cosmic ionizing radiation incidents in aircraft components and crews has been a source of concern and motivated increasingly studies and improvements in the area. The low dose rates involved in this radiation field in aircraft flight altitudes imply Dosimetric necessity of using materials with high efficiency of detection, to enable studies lower cumulative doses resulting in shorter routes or lower altitude. The choice of thermoluminescent dosimeters LiF: Mg, Cu, P was done by having a detection efficiency of about fifteen times higher than its predecessor (LiF: Mg, Ti), and therefore, applied in very low doses dosimetry, and environmental dosimetry . The implementation of the use of pair dosimetric TLD-600H and 700H-TLD will serve as support for testing and studies on the effects of low doses of cosmic radiation in environmental dosimetry applied in the aviation environment in the usual flight altitudes. In this paper are presented the results of development of a methodology for dosimetry low doses of gamma radiation and neutrons using the pair dosimetric TLD-600H and 700H-TLD. The results demonstrate a sensitivity of dosimeters well above the dosimeters LiF: Mg, Ti confirming its suitability for dosimetry of low doses

  20. Development of portable ESR spectrometer as a reader for alanine dosimeters

    International Nuclear Information System (INIS)

    Kojima, T.; Haruyama, Y.; Tachibana, H.; Tanaka, R.; Okamoto, J.

    1993-01-01

    A prototype portable electron spin resonance (ESR) spectrometer was designed and tested, and its feasibility as a reader of alanine dosimeters was studied from the two standpoints of reproducibility of readings and sensitivity sufficient for dosimetry in the absorbed dose range 1-100 kGy. It has two main components: a permanent magnet and resonator; and a unit box with a microwave and auto-frequency control (AFC) circuit, a sweep controller of magnetic field, display, etc. In the present preliminary study, reproducibility values are measured with the same ESR parameters and alanine-polystyrene (alanine-PS) dosimeter at a dose of 1 kGy: repeatedly measuring without removing dosimeter from the cavity; individual measurement with removing and inserting again into the cavity with readjustment of ESR parameters. Alanine/ESR dosimetry using this spectrometer has a measurable dose range from 1 to 100 kGy with relatively high precision within ± 3% (1σ) as a preliminary result. The portable ESR spectrometer may also be modified as an automatic, more precise, dedicated alanine dosimeter reader. (author)

  1. NRC TLD Direct Radiation Monitoring Network

    International Nuclear Information System (INIS)

    Struckmeyer, R.; McNamara, N.

    1991-04-01

    This report presents the results of the NRC [Nuclear Regulatory Commission] Direct Radiation Monitoring Network for the fourth quarter of 1990. It provides the ambient radiation levels measured in the vicinity of 75 sites throughout the United States. In addition, it describes the equipment used, monitoring station selection criteria, characterization of the dosimeter response, calibration procedures, statistical methods, intercomparison, and quality assurance program. 3 figs., 4 tabs

  2. Groundwater Monitoring Plan for the Z-Area Saltstone Facility

    International Nuclear Information System (INIS)

    Wells, D.

    2002-01-01

    Groundwater monitoring has been conducted at the Z-Area Saltstone Disposal Facility since 1987. At that time, groundwater monitoring was not required by the industrial landfill regulations, but a modest monitoring program was required by the operating permit. In 1996 SRS proposed a program based on direct push sampling. This program called for biennial direct push sampling within 25 feet of each waste-containing cell with additional samples being taken in areas where excessive cracking had been observed. The direct push proposal was accepted by The South Carolina Department of Health and Environmental Control (SCDHEC). The Industrial Solid Waste Landfill Regulations were revised in 1998 and now include requirements for groundwater monitoring. The major elements of those regulations and their application at Z-Area are discussed. These are a point of compliance, groundwater protection standards, the groundwater monitoring system, sampling and analysis, and data evaluation and reporting

  3. Survey monitoring of environmental radioactivity in Daejeon area

    Energy Technology Data Exchange (ETDEWEB)

    Jun, Jae Shik.; Noh, Hyung Ah [Taejon Radiation Monitoring Station, Daejeon (Korea, Republic of)

    2000-12-15

    Systematic understanding of the distribution of environmental radioactivity and radiation level in Taejon, including Chungchung area, in normal circumstance, and rapid detection of unusual variation of the radiation level in emergency situation thereby ensure public safety are the objectives of this project to be carried out. This report summarizes and interprets environmental radiation/radioactivity monitoring data obtained at Taejon Radiation Monitoring Station in 2000. In conclusion, the natural environmental radiation level in Taejon area has been preserved as usual and no significant artificial enhancement in environmental radioactivity was observed during the course of this survey and monitoring period.

  4. Survey monitoring of environmental radioactivity in Daejeon area

    Energy Technology Data Exchange (ETDEWEB)

    Jun, Jae Sik; Noh, Hyung Ah [Daejon Radiation Monitoring Station, Daejeon (Korea, Republic of)

    2001-12-15

    Systematic understanding of the distribution of environmental radioactivity and radiation level in Daejeon, including Chungchung area, in normal circumstance, and rapid detection of unusual variation of the radiation level in emergency situation thereby ensure public safety are the objectives of this project to be carried out. This report summarizes and interprets environmental radiation/radioactivity monitoring data obtained at Daejeon Radiation Monitoring Station in 2001. In conclusion, the natural environmental radiation level in Daejeon area has been preserved as usual and no significant artificial enhancement in environmental radioactivity was observed during the course of this survey and monitoring period.

  5. Survey monitoring of environmental radioactivity in Daejeon area

    Energy Technology Data Exchange (ETDEWEB)

    Jun, Jae Shik; Noh, Hyung Ah [Daejon Radiation Monitoring Station, Daejeon (Korea, Republic of)

    2002-12-15

    Systematic understanding of the distribution of environmental radioactivity and radiation level in Daejeon, including Chungchung area, in normal circumstance, and rapid detection of unusual variation of the radiation level in emergency situation thereby ensure public safety are the objectives of this project to be carried out. This report summarizes and interprets environmental radiation/radioactivity monitoring data obtained at Daejeon Radiation Monitoring Station in 2002. In conclusion, the natural environmental radiation level in Daejeon area has been preserved as usual and no significant artificial enhancement in environmental radioactivity was observed during the course of this survey and monitoring period.

  6. Survey monitoring of environmental radioactivity in Daejeon area

    Energy Technology Data Exchange (ETDEWEB)

    Jun, Jae Shik; Noh, Hyung Ah [Daejon Radiation Monitoring Station, Daejeon (Korea, Republic of)

    2003-12-15

    Systematic understanding of the distribution of environmental radioactivity and radiation level in Daejeon, including Chungchung area, in normal circumstance, and rapid detection of unusual variation of the radiation level in emergency situation thereby ensure public safety are the objectives of this project to be carried out. This report summarizes and interprets environmental radiation/radioactivity monitoring data obtained at Daejeon Radiation Monitoring Station in 2002. In conclusion, the natural environmental radiation level in Daejeon area has been preserved as usual and no significant artificial enhancement in environmental radioactivity was observed during the course of this survey and monitoring period.

  7. Survey monitoring of environmental radioactivity in Daejeon area

    International Nuclear Information System (INIS)

    Jun, J. S.; Kim, K. S.

    1998-01-01

    Systematic understanding of the distribution of environmental radioactivity and radiation level in Taejon, including Chungchung area, in normal circumstance, and rapid detection of unusual variation of the radiation level in emergency situation thereby ensure public safety are the objectives of this project to be carried out. This report summarizes and interprets environmental radiation/radioactivity monitoring data obtained at Taejon Radiation Monitoring Station in 1998. In conclusion, the natural environmental radiation level in Taejon area has been preserved as usual and no significant artificial enhancement in environmental radioactivity was observed during the course of this survey and monitoring period

  8. Survey monitoring of environmental radioactivity in Daejeon area

    International Nuclear Information System (INIS)

    Jun, Jae Sik; Noh, Hyung Ah

    2001-12-01

    Systematic understanding of the distribution of environmental radioactivity and radiation level in Daejeon, including Chungchung area, in normal circumstance, and rapid detection of unusual variation of the radiation level in emergency situation thereby ensure public safety are the objectives of this project to be carried out. This report summarizes and interprets environmental radiation/radioactivity monitoring data obtained at Daejeon Radiation Monitoring Station in 2001. In conclusion, the natural environmental radiation level in Daejeon area has been preserved as usual and no significant artificial enhancement in environmental radioactivity was observed during the course of this survey and monitoring period

  9. Direct and pulsed current annealing of p-MOSFET based dosimeter: the "MOSkin".

    Science.gov (United States)

    Alshaikh, Sami; Carolan, Martin; Petasecca, Marco; Lerch, Michael; Metcalfe, Peter; Rosenfeld, Anatoly

    2014-06-01

    Contemporary radiation therapy (RT) is complicated and requires sophisticated real-time quality assurance (QA). While 3D real-time dosimetry is most preferable in RT, it is currently not fully realised. A small, easy to use and inexpensive point dosimeter with real-time and in vivo capabilities is an option for routine QA. Such a dosimeter is essential for skin, in vivo or interface dosimetry in phantoms for treatment plan verification. The metal-oxide-semiconductor-field-effect-transistor (MOSFET) detector is one of the best choices for these purposes, however, the MOSFETs sensitivity and its signal stability degrade after essential irradiation which limits its lifespan. The accumulation of positive charge on the gate oxide and the creation of interface traps near the silicon-silicon dioxide layer is the primary physical phenomena responsible for this degradation. The aim of this study is to investigate MOSFET dosimeter recovery using two proposed annealing techniques: direct current (DC) and pulsed current (PC), both based on hot charged carrier injection into the gate oxide of the p-MOSFET dosimeter. The investigated MOSFETs were reused multiple times using an irradiation-annealing cycle. The effect of the current-annealing parameters was investigated for the dosimetric characteristics of the recovered MOSFET dosimeters such as linearity, sensitivity and initial threshold voltage. Both annealing techniques demonstrated excellent results in terms of maintaining a stable response, linearity and sensitivity of the MOSFET dosimeter. However, PC annealing is more preferable than DC annealing as it offers better dose response linearity of the reused MOSFET and has a very short annealing time.

  10. An ESR study on biological dosimeters: Human hair

    International Nuclear Information System (INIS)

    Colak, Seyda; Ozbey, Turan

    2011-01-01

    In the present work, characteristic features of the radicals found in untreated, gamma and UV-irradiated and mechanical damaged human hair samples were investigated by ESR spectroscopy. Heights of the resonance peaks measured with respect to the spectrum base line were used to monitor microwave power, dose-response, storage time and temperature dependent kinetic features of the radical species contributing to the formation of recorded experimental ESR spectra. Peak heights and g-values (2.0037-2.0052) determined from recorded spectra of hair were color dependent with ΔHpp-0.47 mT. The act of cutting hair samples gene rates sulfur centered radicals which are found in the a-keratin structure of hair. The variations of the peak heights with temperature were related with the water content found in the hair samples. In the 6-1100 Gy dose range, a linear + quadratic dose-response curve was recorded for hair and the mean radiation yield (G mean ) was calculated to be 0.4. The gamma radiation induced radicals were stable for a several hours at room temperature storage conditions. Based on these findings it was concluded that human hair samples could be used as biological/personnel dosimeters and that ESR spectroscopy could be successfully used as a potential technique for monitoring its dosimetric behaviours.

  11. Nevada Test Site 2000 Waste Management Monitoring Report Area 3 and Area 5 Radioactive Waste Management Sites

    International Nuclear Information System (INIS)

    Yvonne Townsend

    2001-01-01

    Environmental monitoring data, subsidence monitoring data, and meteorology monitoring data were collected at and around the Area 3 and Area 5 Radioactive Waste Management Sites (RWMSs) at the Nevada Test Site (NTS) (refer to Figure 1). These monitoring data include radiation exposure, air, groundwater, meteorology, vadose zone, subsidence, and biota data. Although some of these media (radiation exposure, air, and groundwater) are reported in detail in other Bechtel Nevada reports (Annual Site Environmental Report [ASER], the National Emissions Standard for Hazardous Air Pollutants [NESHAP] report, and the Annual Groundwater Monitoring Report), they are also summarized in this report to provide an overall evaluation of RWMS performance and environmental compliance. Direct radiation monitoring data indicate that exposure at and around the RWMSs is not above background levels. Air monitoring data indicate that tritium concentrations are slightly above background levels, whereas radon concentrations are not above background levels. Groundwater monitoring data indicate that the groundwater in the uppermost aquifer beneath the Area 5 RWMS has not been affected by the facility. Meteorology data indicate that 2000 was an average rainfall year: rainfall totaled 167 mm (6.6 in) at the Area 3 RWMS (annual average is 156 mm [6.5 in]) and 123 mm (4.8 in) at the Area 5 RWMS (annual average is 127 mm [5.0 in]). Vadose zone monitoring data indicate that 2000 rainfall infiltrated less than one meter (3 ft) before being returned to the atmosphere by evaporation. Soil-gas tritium monitoring data indicate slow subsurface migration, and tritium concentrations in biota were lower than in previous years. All 2000 monitoring data indicate that the Area 3 and Area 5 RWMSs are performing well at isolating buried waste

  12. Water equivalence of NIPAM based polymer gel dosimeters with enhanced sensitivity for x-ray CT

    International Nuclear Information System (INIS)

    Gorjiara, Tina; Hill, Robin; Bosi, Stephen; Kuncic, Zdenka; Baldock, Clive

    2013-01-01

    Two new formulations of N-isopropylacrylamide (NIPAM) based three dimensional (3D) gel dosimeters have recently been developed with improved sensitivity to x-ray CT readout, one without any co-solvent and the other one with isopropanol co-solvent. The water equivalence of the NIPAM gel dosimeters was investigated using different methods to calculate their radiological properties including: density, electron density, number of electrons per grams, effective atomic number, photon interaction probabilities, mass attenuation and energy absorption coefficients, electron collisional, radiative and total mass stopping powers and electron mass scattering power. Monte Carlo modelling was also used to compare the dose response of these gel dosimeters with water for kilovoltage and megavoltage x-ray beams and for megavoltage electron beams. We found that the density and electron density of the co-solvent free gel dosimeter are more water equivalent with less than a 2.6% difference compared to a 5.7% difference for the isopropanol gel dosimeter. Both the co-solvent free and isopropanol solvent gel dosimeters have lower effective atomic numbers than water, differing by 2.2% and 6.5%, respectively. As a result, their photoelectric absorption interaction probabilities are up to 6% and 19% different from water, respectively. Compton scattering and pair production interaction probabilities of NIPAM gel with isopropanol differ by up to 10% from water while for the co-solvent free gel, the differences are 3%. Mass attenuation and energy absorption coefficients of the co-solvent free gel dosimeter and the isopropanol gel dosimeter are up to 7% and 19% lower than water, respectively. Collisional and total mass stopping powers of both gel dosimeters differ by less than 2% from those of water. The dose response of the co-solvent free gel dosimeter is water equivalent (with 100 keV, correction factor is required for the gels. • For MV electron, correction factor needed for the gels to

  13. Wide area monitoring study

    International Nuclear Information System (INIS)

    Wogman, N.A.; Holdren, G.R. Jr.

    1999-01-01

    Environmental sampling can be used to complement the safeguarding of nuclear material, especially in the detection of undeclared nuclear activities. Routine monitoring of nuclear installations has provided valuable information about the fate of key signature materials within different environmental settings. The approach collates information regarding the generation of individual radiochemical signatures within different nuclear processes, the potential for release of these signatures to the environment and, the chemical form and mobility of the signatures in environmental media along which the material could migrate. Meteorological, geological and hydrological information is used to determine where to sample, what to sample, and how often to sample to provide the greatest likelihood for detection. Multiple strategies can be used to implement wide area monitoring for safeguards purposes. The most complex, and expensive of these, involves establishing extensive networks of fixed location sampling sites. The sites would be operated continuously, and would be instrumented with automated sampling, analysis, and communication equipment to relay information regarding potential anomalies to control centers in near-real time. Alternative strategies can be used to supplement fixed location monitoring equipment, especially in regions that cannot support (financially or logistically) the fixed stations. Through combinations of these various strategies, using a variety of environmental media to monitor a region, we believe that a competent network, one with a quantifiable probability for detecting undeclared nuclear activities, can be designed. While this approach cannot and should not replace other inspection and monitoring activities, it can potentially contribute valuable information to an international safeguards system. (author)

  14. Calibration of gamma cell 220 excel irradiator using Fricke and alanine dosimeters

    International Nuclear Information System (INIS)

    Rushdi, M. A. H.

    2006-06-01

    Using of gamma cell 220 excel irradiators is widely spread in many countries. This type of irradiators is being used for research purposes. Gamma cell 220 excel was provided by the International Atomic Energy Agency (IAEA) to the radiation processing laboratory of Sudan Atomic Energy Commission (SAEC). It is a self-contained gamma irradiator and self shielded, this makes it operates safely. Dose calibration for this cell is important for samples irradiation. In this work, a dosimetry system for the GC220E of SAEC was established using Fricke dosimeter. Fricke dosimeter has a confidence 95% in the rang not exceed 400 Gy. To establish routine dosimetry at high doses up to 5000 Gy, alanine dosimeter was used. This range can demonstrate the ability of GC220E to deliver known controllable doses in reproducible manner for high doses. The irradiation specifications often include a lower and upper limit of absorbed dose or central target dose. Absorbed dose mapping was carried out by both dosimeters to determine the magnitude and locations of D.max and D.min in the irradiation chamber. The results are in good agreement with dose distribution given in the machine manual. A comparison between the tow dosimeters was done and explained.(Author)

  15. Towards a new thickness-independent gamma radiation plastic film dosimeter

    International Nuclear Information System (INIS)

    Vieira, Marli Barbosa; Araujo, Patricia L.; Araujo, Elma S.

    2013-01-01

    A 100% national single-use gamma radiation plastic film dosimeter is presented in this work. A new approach for the development of this material allowed a step forward in the performance of poly (methyl metacrylate) films (PMMA) colored with bromothymol blue (BTB) acid-base indicator. We manage to improve dosimeter performance by introducing a gamma radiation insensitive dye to compensate film thickness variations. By doing so, we were able to obtain consistent dose-response correlations within a set of samples presenting 46 to 110 micrometers in thickness. Hence, our PMMA/BTB-P film dosimeter is suitable to measure absorbed dose in the 2-100kGy range even when film thickness undergoes more than 100% of variation. In addition, dose response data remain practically unaltered for four months after the exposure, when dosimeter films are kept in dark conditions and under refrigeration. The radiation effects on the optical properties were evaluated for Ultraviolet-Visible (UV-Vis) spectrophotometric analysis. Data of characteristic dose-response correlation in terms of changes in the maximum UV-Vis absorption due to radiation, and stability in time are also described. This potential new product is a promising tool for industrial radiation facilities, especially in gamma sterilization of medical supplies. (author)

  16. Acceptance Testing of Thermoluminescent Dosimeter Holders.

    Science.gov (United States)

    Romanyukha, Alexander; Grypp, Matthew D; Sharp, Thad J; DiRito, John N; Nelson, Martin E; Mavrogianis, Stanley T; Torres, Jeancarlo; Benevides, Luis A

    2018-05-01

    The U.S. Navy uses the Harshaw 8840/8841 dosimetric (DT-702/PD) system, which employs LiF:Mg,Cu,P thermoluminescent dosimeters (TLDs), developed and produced by Thermo Fisher Scientific (TFS). The dosimeter consists of four LiF:Mg,Cu,P elements, mounted in Teflon® on an aluminum card and placed in a plastic holder. The holder contains a unique filter for each chip made of copper, acrylonitrile butadiene styrene (ABS), Mylar®, and tin. For accredited dosimetry labs, the ISO/IEC 17025:2005(E) requires an acceptance procedure for all new equipment. The Naval Dosimetry Center (NDC) has developed and tested a new non-destructive procedure, which enables the verification and the evaluation of embedded filters in the holders. Testing is based on attenuation measurements of low-energy radiation transmitted through each filter in a representative sample group of holders to verify that the correct filter type and thickness are present. The measured response ratios are then compared with the expected response ratios. In addition, each element's measured response is compared to the mean response of the group. The test was designed and tested to identify significant nonconformities, such as missing copper or tin filters, double copper or double tin filters, or other nonconformities that may impact TLD response ratios. During the implementation of the developed procedure, testing revealed a holder with a double copper filter. To complete the evaluation, the impact of the nonconformities on proficiency testing was examined. The evaluation revealed failures in proficiency testing categories III and IV when these dosimeters were irradiated to high-energy betas.

  17. The passive radon-thoron discriminative dosimeter for practical use

    International Nuclear Information System (INIS)

    Doi, Masahiro; Kobayashi, Sadayoshi

    1994-01-01

    A passive radon-thoron discriminative dosimeter for practical use has been developed. The body of the practical R-T dosimeter is made of two hemispheric diffusion chambers of carbonized plastic whose diameters are 110 mm and 70 mm, respectively. These diameters are determined to improve the detection efficiency of radon as well as thoron and also the discrimination ratio of radon to thoron. Inner surface of the detector housing is smooth and free from electrified charge to assure the uniform deposition of radon and thoron progeny, because the detector housing is molded out of carbonized plastic as an anti-static material. In addition, structure of an air inlet has improved to contact more tightly with a glass fiber filter to prevent dust from entering the detector housing. The air inlet of the detector housing is also covered with a half-cutted hemispherical windbreak to protect the glass fiber filter from weathering and to stabilize the influence of convectional air flow on the radon and thoron entry rate into two hemispherical diffusion chambers of the dosimeter. The results of calibration exercises showed that the lower detection limit of radon and thoron concentrations were estimated to be 5.1 Bqm -3 and 7.9 Bqm -3 respectively in 2 months exposure. And an interim measurement in the concrete cellar proved that the practical R-T dosimeter has enough specifications to be used in the large-scale radon-thoron discriminative survey. (author)

  18. Medical extrapolation chamber dosimeter model XW6012A

    International Nuclear Information System (INIS)

    Jin Tao; Wang Mi; Wu Jinzheng; Guo Qi

    1992-01-01

    An extrapolation chamber dosimeter has been developed for clinical dosimetry of electron beams and X-rays from medical linear accelerators. It consists of a new type extrapolation chamber, a water phantom and an intelligent portable instrument. With a thin entrance window and a φ20 mm collecting electrode made of polystyrene, the electrode spacing can be varied from 0.2 to 6 mm. The dosimeter can accomplish dose measurement automatically, and has functions of error self-diagnosis and dose self-recording. The energy range applicable is 0.5-20 MeV, and the dose-rate range 0.02-40 Gy/min. The total uncertainty is 2.7%

  19. Performance evaluation of a colorimetric hydrazine dosimeter

    Science.gov (United States)

    Brenner, Karen P.; Rose-Pehrsson, Susan L.

    1994-06-01

    A dosimeter for real-time, colorimetric detection of hydrazine in air has been developed. The passive badge consists of a dosimeter card containing a vanillin solution coated on a thin paper substrate. The active patch consists of a thick cellulose substrate coated with a vanillin solution. When placed in a plastic sample holder attached to a personnel pump, up to 5 L/min can be drawn through the active badge substrate. Through a condensation reaction, vanillin reacts with hydrazine to form a colored product that absorbs in the visible region. The hydrazone formed in the reaction is yellow; its intensity is proportional to the dose. When exposed passively to hydrazine, the experimental detection limit is less than 20 ppb-hrs. Extrapolated results indicate a detection limit of less than 5 ppb-hrs for long sampling periods. Actively sampling of hydrazine vapors gives an experimental detection limit of less than 100 ppb-L at a sample rate of 5 L/min. Relative humidity effects on badge response were minor. High humidity enhanced the color development on the vanillin badge; while low humidity had no effect on badge response. Interference testing of the dosimeters revealed a tobacco smoke interference. Preliminary shelf life tests indicated no decrease in sensitivity to hydrazine when stored at room temperature for 6 months.

  20. Implementation of a remote system for monitoring of radiological areas of radiological areas

    International Nuclear Information System (INIS)

    Velazquez E, Walter; Galuppo G, Emiliano; Gutierrez G, Jorge; Reyes R, Jerson

    2008-01-01

    Full text: Introduction: The present work shows the development of a radiation remote monitoring system which control radiological areas in the principal facilities at CCHEN and the development in the last years to use this system called SMARR (Remote Radiological Area Monitoring System). This is an important issue in radiological safety is to know 'on line' and in a 'continuously way' the radiological variables of areas, especially if in these areas people manage radioactive sources or material, the monitoring system are operative on La Reina and Lo Aguirre Nuclear Centers. This 'knowledge' gets a good support to the radiological safety to safeguard the environment and people in the facilities. Nuclear Chilean Commission: Actually, this system is daily operating to register the background radiation and level operation, for example of the facilities research reactor, cyclone, irradiators, in order to probe the behaviors under operational requirements. The system was made using common Geiger Muller and NaI detectors. This signal is received, data by data, for a collector computer which uses a Labview program to do this displayed on a screen computer using graphics to show the activity on a radiological area, and when the lectures pass a setting value automatically the system send by e-mail and text message which also can be received for cell phones enabled for this for the supervisor. Each monitored facility is completely independent of each other and store a data backup, also every installation are monitoring with server computer, it's concentrating the information and allow to view it on line in real time, trough the intranet and internet network. In addition, the information is stored in the special report in the server and available for to do a statistics and identify the operation periods, and control of radioactive sources. The Industry: The radiological protection on industry is necessary today, the typical instrumentation on the industry is growing up in the

  1. Performance characteristics of a microMOSFET as an in vivo dosimeter in radiation therapy

    International Nuclear Information System (INIS)

    Ramaseshan, R; Kohli, K S; Zhang, T J; Lam, T; Norlinger, B; Hallil, A; Islam, M

    2004-01-01

    The commercially available microMOSFET dosimeter was characterized for its dosimetric properties in radiotherapy treatments. The MOSFET exhibited excellent correlation with the dose and was linear in the range of 5-500 cGy. No measurable effect in response was observed in the temperature range of 20-40 0 C. No significant change in response was observed by changing the dose rate between 100 and 600 monitor units (MU) min -1 or change in the dose per pulse. A 3% post-irradiation fading was observed within the first 5 h of exposure and thereafter it remained stable up to 60 h. A uniform energy response was observed in the therapy range between 4 MV and 18 MV. However, below 0.6 MeV (Cs-132), the MOSFET response increased with the decrease in energy. The MOSFET also had a uniform dose response in 6-20 MeV electron beams. The directional dependence of MOSFET was within ±2% for all the energies studied. The inherent build-up of the MOSFET was evaluated dosimetrically and found to have varying water equivalent thickness, depending on the energy and the side of the beam entry. At depth, a single calibration factor obtained by averaging the MOSFET response over different field sizes, energies, orientation and depths reproduced the ion chamber measured dose to within 5%. The stereotactic and the penumbral measurements demonstrated that the MOSFET could be used in a high gradient field such as IMRT. The study showed that the microMOSFET dosimeter could be used as an in vivo dosimeter to verify the dose delivery to the patient to within ±5%

  2. Quo vadis, personnel monitoring

    International Nuclear Information System (INIS)

    Becker, K.

    1975-01-01

    With the increasing use of nuclear power and radiation sources, the selection of optimum systems for personnel monitoring is becoming a matter of worldwide concern. The present status of personnel dosimetry, sometimes characterized by unstable and inaccurate detectors and oversimplified interpretation of the results, leaves much to be desired. In particular, photographic film, although having certain advantages with regard to economics and information content, undergoes rapid changes in warm and humid climates. Careful sealing reduces, but does not prevent, these problems. The replacement of film by solid-state dosimeters, primarily thermoluminescence dosimeters, is in progress or being considered by an increasing number of institutions and requires a number of decisions concerning the choice of the optimum detector(s), badge design, and evaluation system; organizational matters, such as the desirability of automation and computerized bookkeeping; etc. The change also implies the potential use of such advanced concepts as different detectors and monitoring periods for the large number of low-risk persons and the small number of high-risk radiation workers. (auth)

  3. Miniature Active Space Radiation Dosimeter, Phase II

    Data.gov (United States)

    National Aeronautics and Space Administration — Space Micro will extend our Phase I R&D to develop a family of miniature, active space radiation dosimeters/particle counters, with a focus on biological/manned...

  4. Argentine Republic intercomparison programme for personal dosimeters

    International Nuclear Information System (INIS)

    Gregori, Beatriz N.; Papadopulos, Susana B.; Kunst, Juan J.; Cruzate, Juan A.; Saravi, Margarita C.

    2004-01-01

    In 1997 an Intercomparison Program for individual monitoring started in order to test (on a voluntary basis) the performance in absorbed dose and personal dose equivalent determinations. The aim of the program was also to gain some insight into the general aspects related to the type of the personnel dosimeter used, the calibration procedures, the phantom spectral dependence and the management of radiological quantities. The Regional Reference Center for Dosimetry (CCR), of the Argentine National Atomic Energy Commission and the Physical Dosimetry Laboratory of the Argentine National Regulatory Authority, performed the irradiations. Those were done free air and on ICRU phantom, using X-ray, quality ISO: W60, W80, W110 and W200; and 137 Cs and 60 Co gamma rays, normal and angular (0, 30, 60 degrees) incidence. In the framework of the Program, an upgraded workshop took place and the national standard, IRAM 17146, was elaborated as well. In this work, the laboratories performance and its temporal evolution is shown from 1997 up to 2002. The suggestions to improve their performance are also included. (author)

  5. Technical Basis Document for PFP Area Monitoring Dosimetry Program

    CERN Document Server

    Cooper, J R

    2000-01-01

    This document describes the phantom dosimetry used for the PFP Area Monitoring program and establishes the basis for the Plutonium Finishing Plant's (PFP) area monitoring dosimetry program in accordance with the following requirements: Title 10, Code of Federal Regulations (CFR), part 835, ''Occupational Radiation Protection'' Part 835.403; Hanford Site Radiological Control Manual (HSRCM-1), Part 514; HNF-PRO-382, Area Dosimetry Program; and PNL-MA-842, Hanford External Dosimetry Technical Basis Manual.

  6. Technical Basis Document for PFP Area Monitoring Dosimetry Program

    International Nuclear Information System (INIS)

    COOPER, J.R.

    2000-01-01

    This document describes the phantom dosimetry used for the PFP Area Monitoring program and establishes the basis for the Plutonium Finishing Plant's (PFP) area monitoring dosimetry program in accordance with the following requirements: Title 10, Code of Federal Regulations (CFR), part 835, ''Occupational Radiation Protection'' Part 835.403; Hanford Site Radiological Control Manual (HSRCM-1), Part 514; HNF-PRO-382, Area Dosimetry Program; and PNL-MA-842, Hanford External Dosimetry Technical Basis Manual

  7. Development of based on 89S51 single-chip microcomputer electronic dosimeter

    International Nuclear Information System (INIS)

    Wang Junhua; Zhou Jiachao; Sun Jianghan; Du Xiao

    2009-01-01

    It describes the main design features and basic properties of based on 89S51 single-chip microcomputer electronic dosimeter with wide range and multi purposes. The dosimeter can display dose rate or accumulative dose or the maximum dose rate, record accumulative dose, the maximum dose rate and classes. (authors)

  8. Small is beautiful: SAIC's new dosimeter

    International Nuclear Information System (INIS)

    Benson, R.G.

    1991-01-01

    Science Applications International Corporation (California) has developed an energy-compensated Geiger tube in a package the size of a small pocket pager. In fact, the whole dosimeter measures just 48mm x 72mm x 17mm. The rugged, lightweight unit is sensitive enough to record radiation ranging from low background levels caused by the earth's surface, the sun, or cosmic radiation, to beyond lethal dose levels. The PD-1 provides dose measurement, dose rate measurement, and ''chip'' functions. A chirper sounds each time a specified dose is accumulated, and the chirp increments are defined by the user. A dosimeter reader provides a simple interface for bi-directional communication with host PC. The Geiger tube provides improved accuracy over a wider energy range than current solid state devices. Features such as long battery life, long calibration life (two years or longer), and easy calibration procedure should help to simplify the work of health physicists overseeing dosimetry management programmes. (author)

  9. Human hair as a pollutant dosimeter

    International Nuclear Information System (INIS)

    Al-Hashimi, A.

    1991-01-01

    Human hair has been proved to be a better dosimeter than even blood for tracing most of the heavy metal toxins when they penetrate the biosphere. The high precision of the neutron activation analysis (NAA) enabled researchers to elegantly differentiate between endogenous and exogenous contamination and thoroughly study poisonings caused by these physiologically-unimportant elements. Extensive volume of bench-scale work has been accomplished in these laboratories to show the capacity of INAA to detect the presence of 10 nuclides (or more) with a precision of about 5%. The principal objective of the present study is to employ this assaying power and the tendency of scalp hair to uptake heavy metals from aqueous solutions, to design a dosimeter which can easily be used by the environmentalists. The findings should also be of interest to the waste-management people who are searching for a cost-effective technique to remove these pollutants from relatively large volumes of industrial effluents

  10. Electrochemical development of particle tracks in CR-39 polymer dosimeter

    International Nuclear Information System (INIS)

    Hadlock, D.E.; Parkhurst, M.A.; Yang, C.S.; Groeger, J.; Johnson, J.R.; Huang, S.J.

    1985-09-01

    Electrochemical etching of CR-39 polymeric track etch neutron detectors results in proton-recoil tracks can be distinguished from background tracks much better than tracks developed solely by chemical etching. A newly designed and constructed electrochemical etching apparatus allows large numbers of dosimeters to be processed simultaneously with consistent results. Many processing systems have been developed for chemical and electrochemical etching of the track etch dosimeters. Three systems specifically show great promise and are being studied extensively

  11. PRESAGE® as a solid 3-D radiation dosimeter: A review article

    International Nuclear Information System (INIS)

    Khezerloo, Davood; Nedaie, Hassan Ali; Takavar, Abbas; Zirak, Alireza; Farhood, Bagher; Movahedinejhad, Hadi; Banaee, Nooshin; Ahmadalidokht, Isa; Knuap, Courtney

    2017-01-01

    Radiation oncology has been rapidly improved by the application of new equipment and techniques. With the advent of new complex and precise radiotherapy techniques such as intensity modulated radiotherapy, stereotactic radiosurgery, and volumetric modulated arc therapy, the demand for an accurate and feasible three-dimensional (3-D) dosimetry system has increased. The most important features of a 3-D dosimeter, apart from being precise, accurate and reproducible, include also its low cost, feasibility, and availability. In 2004 a new generation of solid plastic dosimeters which demonstrate a radiochromic response to ionizing radiation was introduced. PRESAGE ® plastic dosimeter lacks the limitations of previous Ferric and polymer plastic 3-D dosimeters such as diffusion, sensitivity to oxygen, fabrication problems, scanning and read out challenges. In this decade, a large number of efforts have been carried out to enhance PRESAGE ® structure and scanning methods. This article attempts to review and reflect on the results of these investigations. - Highlights: • Sensitivity and stability can improve with variation in weight fraction of gel. • To overcome star and edge artifacts, Wide-parallel beam optical CT can use in clinic. • Modeling of scatter pattern can be usable to enhance of images.

  12. Nevada Test Site 2001 Waste Management Monitoring Report Area 3 and Area 5 Radioactive Waste Management Sites; TOPICAL

    International Nuclear Information System (INIS)

    Y. E. Townsend

    2002-01-01

    Environmental monitoring data, subsidence monitoring data, and meteorology monitoring data were collected at and around the Area 3 and Area 5 Radioactive Waste Management Sites (RWMSs) at the Nevada Test Site (NTS) (refer to Figure 1). These monitoring data include radiation exposure, air, groundwater, meteorology, vadose zone, subsidence, and biota data. Although some of these media (radiation exposure, air, and groundwater) are reported in detail in other Bechtel Nevada (BN) reports (Annual Site Environmental Report[ASER], the National Emissions Standard for Hazardous Air Pollutants[NESHAP] report, and the Annual Groundwater Monitoring Report), they are also summarized in this report to provide an overall evaluation of RWMS performance and environmental compliance. Direct radiation monitoring data indicate that exposure at and around the RWMSs is not above background levels. Air monitoring data indicate that tritium concentrations are slightly above background levels. Groundwater monitoring data indicate that the groundwater in the uppermost aquifer beneath the Area 5 RWMS has not been affected by the facility. Meteorology data indicate that 2001 was an average rainfall year: rainfall totaled 150 mm (5.9 in) at the Area 3 RWMS and 120 mm (4.7 in) at the Area 5 RWMS. Vadose zone monitoring data indicate that 2001 rainfall infiltrated less than one meter (3 ft) before being returned to the atmosphere by evaporation. Soil-gas tritium monitoring data indicate slow subsurface migration, and tritium concentrations in biota were lower than in previous years. All 2001 monitoring data indicate that the Area 3 and Area 5 RWMSs are performing within expectations of the model and parameter assumptions for the facility performance assessments

  13. Water equivalency evaluation of PRESAGE® dosimeters for dosimetry of Cs-137 and Ir-192 brachytherapy sources

    Science.gov (United States)

    Gorjiara, Tina; Hill, Robin; Kuncic, Zdenka; Baldock, Clive

    2010-11-01

    A major challenge in brachytherapy dosimetry is the measurement of steep dose gradients. This can be achieved with a high spatial resolution three dimensional (3D) dosimeter. PRESAGE® is a polyurethane based dosimeter which is suitable for 3D dosimetry. Since an ideal dosimeter is radiologically water equivalent, we have investigated the relative dose response of three different PRESAGE® formulations, two with a lower chloride and bromide content than original one, for Cs-137 and Ir-192 brachytherapy sources. Doses were calculated using the EGSnrc Monte Carlo package. Our results indicate that PRESAGE® dosimeters are suitable for relative dose measurement of Cs-137 and Ir-192 brachytherapy sources and the lower halogen content PRESAGE® dosimeters are more water equivalent than the original formulation.

  14. Blind test results - measure of incessant progress in personnel monitoring services at Kudankulam Nuclear Power Project

    International Nuclear Information System (INIS)

    Jashi, K.B.; Ashok Kumar, A.; Pandaram, P.

    2018-01-01

    Personnel Monitoring Services (PMS) at Thermoluminescent Dosimeter (TLD) laboratory, Kudankulam Nuclear Power Project (KKNPP) involve monitoring of about 1500 radiation workers of KKNPP for external radiation exposure on monthly basis. Quality assurance in PMS provides an overall estimate of accuracy which depends not only upon the dosimeter quality and reader performance but also on data interpretation, applicable software and the system. Therefore it measures the effectiveness of quality control mechanism. Blind test is an effective tool to check the performance of the processor without their knowledge during routine processing of the service cards

  15. Technical Basis Document for PFP Area Monitoring Dosimetry Program

    Energy Technology Data Exchange (ETDEWEB)

    COOPER, J.R.

    2000-04-17

    This document describes the phantom dosimetry used for the PFP Area Monitoring program and establishes the basis for the Plutonium Finishing Plant's (PFP) area monitoring dosimetry program in accordance with the following requirements: Title 10, Code of Federal Regulations (CFR), part 835, ''Occupational Radiation Protection'' Part 835.403; Hanford Site Radiological Control Manual (HSRCM-1), Part 514; HNF-PRO-382, Area Dosimetry Program; and PNL-MA-842, Hanford External Dosimetry Technical Basis Manual.

  16. An improved design of QFE dosimeter charging unit

    International Nuclear Information System (INIS)

    Speight, R.G.; Clifton, J.J.

    1976-02-01

    The availability of commercial charging units for quartz fibre electroscope dosimeters (QFEs) in the United Kingdom has been limited to small battery units intended for use in laboratory conditions. The use of large numbers of QFEs by semi-skilled staff has resulted in damage to many dosimeters, particularly at the charging pins. Difficulties have also been experienced in the viewing of the scale and fibre in non-laboratory lighting conditions. These problems have resulted in the development of a robust charging and viewing unit, which is described in this report. The new unit which is mains electricity powered, is contained in a case 230 x 305 x 150 mm weighing 4.5 kg. (U.K.)

  17. Design characteristics of a three-component AEOI Neutriran Albedo Neutron Personnel Dosimeter

    International Nuclear Information System (INIS)

    Sohrabi, M.; Katouzi, M.

    1991-01-01

    In establishing a national personnel neutron dosimetry service in Iran, different parameters of the AEOI Neutriran Albedo Neutron Personnel Dosimeter (NANPD) have been optimized. A NANPD was designed with three dosimetry components to measure (a) direct thermal neutrons, (b) direct fast neutrons and (C) direct neutrons by the detection of the albedo neutrons reflected from the body. The dosimeter consists of one or more Lexan polycarbonate and/or CR-39 foils and two 10 B (n,α) 7 Li converters in a cadmium cover so arranged as to efficiently measure the three neutron dose components separately. The boron converter thickness, its position relative to the beam direction and its distance from the PC foil were studied and the results were incorporated into the design. The dose response of the dosimeter, its lower detection limit as well as the correction factors related to the field neutrons and albedo neutrons were also determined for a 238 Pu-Be, an 241 Am-Be and a 252 Cf sources. In this paper, the dosimeter design and its dosimetric characteristics are presented and discussed. (author)

  18. Performance improvement of pentacosa-diynoic acid label dosimeter for radiation processing technology

    Science.gov (United States)

    Abdel-Fattah, A. A.; Soliman, Y. S.

    2017-12-01

    A radiation sensitive material, 10,12-pentacosa-diynoic acid (PCDA), was incorporated into polyvinyl butyral (PVB) films to develop indicators/dosimeters for blood and food irradiation. The present study aims to improve the dosimetric performance of these previously prepared dosimeters and to extend their shelf life by the combination of a radical scavenger, propyl gallate (PG), and a UV absorber, tinuvin-p (TP). The X-ray diffraction (XRD) patterns of the dosimeters were analysed and their dosimetric characteristics were investigated by specular reflectance in the visible spectrum range of 400-700 nm. Upon irradiation, the films turn blue exhibiting two main bands around 670 and 620 nm. Their dose-response functions were fitted by a double exponential growth, 5 parameters, equation. Irradiation temperature influences the dosimeter response at 670 nm without causing thermochromic transition up to 50 °C in poly-PCDA. The useful dose range is 5-4000 Gy depending on the wavelengths of analysis and PCDA content in the films. The overall uncertainty of dose measurement is less than 6% at 2σ.

  19. REMOTE AREA RADIATION MONITORING (RARM) ALTERNATIVES ANALYSIS

    International Nuclear Information System (INIS)

    NELSON RL

    2008-01-01

    The Remote Area Radiation Monitoring (RARM) system will be used to provide real-time radiation monitoring information to the operations personnel during tank retrieval and transfer operations. The primary focus of the system is to detect potential anomalous (waste leaks) or transient radiological conditions. This system will provide mobile, real-time radiological monitoring, data logging, and status at pre-selected strategic points along the waste transfer route during tank retrieval operations. The system will provide early detection and response capabilities for the Retrieval and Closure Operations organization and Radiological Control personnel

  20. Determination of dose correction factor for energy and directional dependence of the MOSFET dosimeter in an anthropomorphic phantom

    International Nuclear Information System (INIS)

    Cho, Sung Koo; Choi, Sang Hyoun; Kim, Chan Hyeong; Na, Seong Ho

    2006-01-01

    In recent years, the MOSFET dosimeter has been widely used in various medical applications such as dose verification in radiation therapeutic and diagnostic applications. The MOSFET dosimeter is, however, mainly made of silicon and shows some energy dependence for low energy photons. Therefore, the MOSFET dosimeter tends to overestimate the dose for low energy scattered photons in a phantom. This study determines the correction factors to compensate these dependences of the MOSFET dosimeter in ATOM phantom. For this, we first constructed a computational model of the ATOM phantom based on the 3D CT image data of the phantom. The voxel phantom was then implemented in a Monte Carlo simulation code and used to calculate the energy spectrum of the photon field at each of the MOSFET dosimeter locations in the phantom. Finally, the correction factors were calculated based on the energy spectrum of the photon field at the dosimeter locations and the pre-determined energy and directional dependence of the MOSFET dosimeter. Our result for 60 Co and 137 Cs photon fields shows that the correction factors are distributed within the range of 0.89 and 0.97 considering all the MOSFET dosimeter locations in the phantom

  1. Single and multichannel scintillating fiber dosimeter for radiotherapic beams with SiPM readout

    Energy Technology Data Exchange (ETDEWEB)

    Berra, A., E-mail: alessandro.berra@gmail.it [Università degli Studi dell' Insubria e INFN sezione di Milano Bicocca (Italy); Ferri, A. [Fondazione Bruno Kessler, Centro per i Materiali e i Microsistemi (Italy); Novati, C. [Università degli Studi dell' Insubria e INFN sezione di Milano Bicocca (Italy); Ostinelli, A. [Ospedale Sant' Anna, Servizio di Fisica Sanitaria (Italy); Paternoster, G.; Piemonte, C. [Fondazione Bruno Kessler, Centro per i Materiali e i Microsistemi (Italy); Prest, M. [Università degli Studi dell' Insubria e INFN sezione di Milano Bicocca (Italy); Vallazza, E. [INFN Sezione di Trieste (Italy)

    2016-12-01

    The treatment of many neoplastic diseases requires the use of radiotherapy, which consists in the irradiation of the tumor, identified as the target volume, with ionizing radiations generated both by administered radiopharmaceuticals or by linear particle accelerators (LINACs). The radiotherapy beam delivered to the patient must be regularly checked to assure the best tumor control probability: this task is performed with dosimeters, i.e. devices able to provide a measurement of the dose deposited in their sensitive volume. This paper describes the development of two scintillator dosimeter prototypes for radiotherapic applications based on plastic scintillating fibers read out by high dynamic range Silicon PhotoMultipliers. The first dosimeter, consisting of a single-channel prototype with a pair of optical fibers, a scintillating and a white one, read out by two SiPMs, has been fully characterized and led to the development of a second multi-channel dosimeter based on an array of scintillating fibers: this device represents the first step towards the assembly of a “one-shot” device, capable to perform some of the daily quality controls in a few seconds. The dosimeters characterization was performed with a Varian Clinac iX linear accelerator at the Radiotherapy Department of the St. Anna Hospital in Como (IT).

  2. Characteristics of a normoxic polymethacrylic acid gel dosimeter for a 72-MeV proton beam

    Science.gov (United States)

    Bong, Jihye; Shin, Dongho; Kwon, Soo-Il

    2014-01-01

    The characteristics of a normoxic polymethacrylic acid gel dosimeter for a 72-MeV proton beam were evaluated. A polymer gel dosimeter was synthesized using gelatin, methacrylic acid, hydroquinone, tetrakis(hydroxymethyl) phosphonium chloride, and highly purified distilled water. The dosimeter was manufactured by placement in a polyethylene (PE) container. Irradiated dosimeters were analyzed to determine the transverse relaxation time (T2) using a 1.5-T MRI. A calibration curve was obtained as a function of the absorbed dose. A Bragg curve made by irradiating the gel with mono-energy was compared with the results for a parallel plate ionization chamber. The spread-out Bragg peak (SOBP) range and distal dose fall-off (DDF) were comparatively analyzed by comparing the irradiated gel with a spread-out Bragg peak against with the ion chamber. Lastly, the gel's usefulness as a dosimeter for therapeutic radiation quality assurance was evaluated by obtaining its practical field size, flatness, and symmetry, through comparison of the profiles of the gel and ion chamber.

  3. Dosimetric characterization of the PTW Seven29 dosimeter and Octavius Phantom for IMRT quality control

    International Nuclear Information System (INIS)

    Goncalves, Leandro R.; Habitzreuter, Angela B.; Santos, Gabriela R.; Watanabe, Erica Y.; Silva, Marco A.; Menegussi, Gisela; Rodrigues, Laura N.; Furnari, Laura

    2012-01-01

    Techniques like IMRT, VMAT and tomotherapy has been used to improve dose conformity in the target, while sparing adjacent normal tissues. The complexity of this techniques challenge to correctly verify the dose delivery, in an independent way. Matrix detectors have been used for this purpose. Although, to exactly understand the dosimeter response and to identify his limitations, characterization measurements need to be performed. These dosimeters, for instance, can present angular dependence. Phantoms has been designed to, when used together the detector, eliminate this angular dependence. The aim of this work was to characterize PTW Seven 29 dosimeter and also his use with Octavius Phantom (PTW). The dosimeter showed reproducible with 0.25% the biggest standard deviation, good dose linearity and dose rate independence. Differences for output factors were obtained (<6%), but a clinical case measurement showed that the set can be used for IMRT verification. When used with Octavius Phantom the dosimeter showed low angular dependence. (author)

  4. Fast neutron dosimetry using CaSO4:Dy thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Azorin, N.G.; Salvi, C.R.; Rubio, J.L.; Gutierrez, C.A.

    1980-01-01

    The use of CaSO 4 :Dy phosphor powder in fast neutron dose measurements using the activation of sulphur from the 32 S(n,p) 32 P reaction is described. The thermoluminescence induced during the irradiation and that due to the decay of short-lived activation products, is erased by annealing the dosimeters after a post-irradiation time of 3 days. The self-induced thermoluminescence measured at different intervals of post-irradiation time, gives an estimation of the fast neutron dose to which the dosimeters were exposed

  5. How does the chemistry of polymer gel dosimeters affect their performance?

    International Nuclear Information System (INIS)

    Jirasek, A; McAuley, K B; Lepage, M

    2009-01-01

    This review article describes the primary chemical processes affecting polymerization in polymer-based radiosensitive gel dosimeters. Furthermore, recently studied environmental factors and gelation processes affecting polymerization and gel dose response are discussed. Finally, the predominant physical factors affecting dose-response sensitivity for each of the three imaging modalities (MRI, Optical CT, x-ray CT) are outlined. Emphasis is concentrated on recent literature in this area, as prior literature was summarized in review articles submitted to DOSGEL 2006 in Sherbrooke (J. Conf. Series, 56, 2006). Discussions from the recent DOSGEL conference (2008) are incorporated into this review.

  6. How does the chemistry of polymer gel dosimeters affect their performance?

    Energy Technology Data Exchange (ETDEWEB)

    Jirasek, A [Department of Physics and Astronomy, University of Victoria, Victoria BC V8W 3P6 (Canada); McAuley, K B [Department of Chemical Engineering, Queen' s University, Kingston ON K7L 3N6 (Canada); Lepage, M [Centre d' imagerie moleculaire de Sherbrooke, Universite de Sherbrooke, 3001 12e Avenue Nord, Sherbrooke QC J1H 5N4 (Canada)], E-mail: jirasek@uvic.ca, E-mail: kim.mcauley@chee.queensu.ca, E-mail: martin.lepage@usherbrooke.ca

    2009-05-01

    This review article describes the primary chemical processes affecting polymerization in polymer-based radiosensitive gel dosimeters. Furthermore, recently studied environmental factors and gelation processes affecting polymerization and gel dose response are discussed. Finally, the predominant physical factors affecting dose-response sensitivity for each of the three imaging modalities (MRI, Optical CT, x-ray CT) are outlined. Emphasis is concentrated on recent literature in this area, as prior literature was summarized in review articles submitted to DOSGEL 2006 in Sherbrooke (J. Conf. Series, 56, 2006). Discussions from the recent DOSGEL conference (2008) are incorporated into this review.

  7. Design and construction of a personal radiation monitor

    International Nuclear Information System (INIS)

    Oliveira, A.H. de.

    1979-01-01

    The design and construction of a personal radiation monitor is dealt with. It provides a direct and reliable reading of the received dose, and sounds an alarm in the event of an excessive dose rate. A detailed analysis of its design, construction and caracteristics is given, as well a comparison with usual dosimeters. The apparatus is based on large-scale integrated electronics, it is reliable, easy to read, and can be used in the same manner as a film-badge. Contrariwise to both the film-badge and the termoluminescent dosimeters, it allows a direct reading. Also, it is not susceptible to false readings due to mechanical shocks, which is the weak point of the pocket ionization chamber. The dosimeter described herein is equipped with a special corrective shield that gives the correct dose reading, regardless of the photon energy. It is also highly immune to neutron interference. (Author) [pt

  8. Application of thermoluminescent dosimeters (CaSO4:Dy) in the evaluation of the dose in ornamental granites of Pernambuco, Brazil

    International Nuclear Information System (INIS)

    Rocha, Edilson Accioly; Cardoso, Jorge J.R.F.; Santos Junior, Jose A. dos; Silva, Cleomacio M. da

    2007-01-01

    The granites of Pernambuco are being used as ornamental rocks, and besides being commercialized in our state. They are also exported to Europe and Asia. The granite is utilized in the fabrication of objects and in wall and floor covering. It is scientifically known because it presents natural series of radionuclides ( 238 U and 232 Th) and 40 K which are higher than the average content of the other rocks. The gamma radiation of the granite coming from the decay of the series cited accompanied by 40 K, must not exceed the limit of 1 mSv.a -1 for the public, according to norms that are in effect. The objective of this work is to evaluate the dose of gamma radiation that is absorbed by individuals who are exposed to the granite. This work is expected to contribute to environmental monitoring and to the evaluation of health risk to man. Pieces of granite from diverse environments were monitored for a total period of 225 days. Three batches of CaSO 4 :Dy were used. The first batch remained fixed to the pieces of granite (one dosimeter for each object) for period of sixty-eight days. After the reading, they were reaffixed to the same objects, each one of them accompanied by one or more dosimeters from the second and third batches. The dosimeters remained affixed for another average period of one hundred and forty-eight days. The absorbed doses found varied from the background up to 0.64 mSv.a -1 , with a general average of 0.24 mSv.a -1 , thus, it was concluded that the doses of granite studied, up to the present, were inside the parameters recommended by CNEN, when considering potentially exposed members of the public. (author)

  9. A gelatin-free model system for the study of the basic radiation-induced polymerization in PAG dosimeters

    International Nuclear Information System (INIS)

    Babic, S; Park, Y S; Schreiner, L J

    2004-01-01

    In this presentation we show results of investigations on gelatin-free dosimeters containing equal amounts of acrylamide and N,N'-methylene-bisacrylamide (named Aqueous Polyacrylamide, APA, dosimeters). The dosimeters were prepared with three different total monomer concentrations (2, 6, and 8% by weight). Nuclear magnetic resonance (NMR) spin-spin and spin-lattice proton relaxation measurements at 20 MHz, and gravimetric analyses performed on all three dosimeters, show a continuous degree of polymerization over the range of dose 0.5 - 25 Gy. The developed NMR model explains the relationship observed between the relaxation data and the amount of cross-linked polymer formed at each dose. This model may be extended with gelatin relaxation data to provide a fundamental understanding of radiation-induced polymerization in the conventional PAG dosimeters

  10. Wide area continuous offender monitoring

    Energy Technology Data Exchange (ETDEWEB)

    Hoshen, J. [Lucent Technologies (United States); Drake, G. [New Mexico Dept. of Corrections, Santa Fe, NM (United States); Spencer, D. [Sandia National Labs., Albuquerque, NM (United States)

    1996-11-01

    The corrections system in the U.S. is supervising over five million offenders. This number is rising fast and so are the direct and indirect costs to society. To improve supervision and reduce the cost of parole and probation, first generation home arrest systems were introduced in 1987. While these systems proved to be helpful to the corrections system, their scope is rather limited because they only cover an offender at a single location and provide only a partial time coverage. To correct the limitations of first-generation systems, second-generation wide area continuous electronic offender monitoring systems, designed to monitor the offender at all times and locations, are now on the drawing board. These systems use radio frequency location technology to track the position of offenders. The challenge for this technology is the development of reliable personal locator devices that are small, lightweight, with long operational battery life, and indoors/outdoors accuracy of 100 meters or less. At the center of a second-generation system is a database that specifies the offender`s home, workplace, commute, and time the offender should be found in each. The database could also define areas from which the offender is excluded. To test compliance, the system would compare the observed coordinates of the offender with the stored location for a given time interval. Database logfiles will also enable law enforcement to determine if a monitored offender was present at a crime scene and thus include or exclude the offender as a potential suspect.

  11. K-band EPR dosimetry: small-field beam profile determination with miniature alanine dosimeter

    International Nuclear Information System (INIS)

    Chen, Felipe; Graeff, Carlos F.O.; Baffa, Oswaldo

    2005-01-01

    The use of small-size alanine dosimeters presents a challenge because the signal intensity is less than the spectrometer sensitivity. K-band (24 GHz) EPR spectrometer seems to be a good compromise between size and sensitivity of the sample. Miniature alanine pellets were evaluated for small-field radiation dosimetry. Dosimeters of DL-alanine/PVC with dimensions of 1.5 mm diameter and 2.5 mm length with 5 mg mass were developed. These dosimeters were irradiated with 10 MV X-rays in the dose range 0.05-60 Gy and the first harmonic (1 h) spectra were recorded. Microwave power, frequency and amplitude of modulation were optimized to obtain the best signal-to-noise ratio (S/N). For beam profile determination, a group of 25 dosimeters were placed in an acrylic device with dimensions of (7.5x2.5x1) cm 3 and irradiated with a (3x3) cm 2 10 MV X-rays beam field size. The dose at the central region of the beam was 20 Gy at a depth of 2.2 cm (build up for acrylic). The acrylic device was oriented perpendicular to the beam axis and to the gantry rotation axis. For the purposes of comparison of the spatial resolution, the beam profile was also determined with a radiographic film and 2 mm aperture optical densitometer; in this case the dose was 1 cGy. The results showed a similar spatial resolution for both types of dosimeters. The dispersion in dose reading was larger for alanine in comparison with the film, but alanine dosimeters can be read faster and more directly than film over a wide dose range

  12. Design and Implementation of Accurate and Efficient Pocket Dosimeter

    International Nuclear Information System (INIS)

    Shehata, S.A.; Abdelkhalek, K.L.

    2005-01-01

    It is so important in the field of radiation therapy and radiation protection to have dosimeters to determine the absorbed dose, which is transferred to human body by ionizing radiation. In this paper the design and implementation of a wide-range pocket dosimeter (PKD-1) with high accuracy to measure personal equivalent dose and dose rate of gamma radiation will be presented. This pocket dosimeter is micro controller-based and powered from 9 V rechargeable battery. The overall power consumption is significantly reduced by smart software and hardware design allowing longer time intervals between recharging. The integrated alphanumerical LCD displays not only of the accumulated dose and current dose rate, but also displays alarm messages such as low battery. For reasons of power saving the LCD is activated on demand by pressing the push button or automatically when an alarm occurs. Audible and visual alarms have been added to PKD-1 in order that they cannot be accidentally overlooked or ignored. PKD-1 can be connected to any PC through its serial port (RS232) and User Interface software has been developed for easy displaying and recording of radiation readings over any time period

  13. Uses of polymer-alanine film/ESR dosimeters in dosimetry of ionizing radiation

    International Nuclear Information System (INIS)

    Xie Liqing; Zhang Yinfeng; Dai Jinxian; Lu Ting; Chen Ruyi; Yang Hua

    1993-01-01

    Alanine ESR dosimetry is a reliable method, used in a various fields of ionizing radiation. The polymer-alanine film/ESR dosimeters of 0.3 -0.4 mm thickness were prepared and their dosimetric properties were studied for 60 Co γ photons and 3 - 5 MeV electrons in the dose range from 20 Gy to 100 kGy. The results show that under normal conditions the alanine calibration curves are linear in the dose range from 100 Gy to 10kGy. The dose profiles at the electron radiation field were measured with the film alanine dosimeters. The polymer-alanine film dosimeters were used for ion implantation of 400 keV ion implantor. Their dose response and energy dependence were investigated initially. (Author)

  14. Optimization of the obtaining method of CaSO4: Dy + Ptfe dosimeters

    International Nuclear Information System (INIS)

    Galicia A, J.; Rioja Ch, J.; Torijano C, E.; Azorin N, J.

    2012-10-01

    This work contain the obtained results of studying the response when irradiating to different dose of X-rays, dosimeters of CaSO 4 : Dy + Ptfe using different lapses of time in their preparation (a lot of selected dosimeters of an elaboration process of 3 days and another of an elaboration process of 2 hours). For the elaboration of the powdered material, the evaporation method was used; the irradiation were carried out in a lineal accelerator Elekta Synergy property of the National Medical Center, 20 de November. The similarities and differences are shown among the two dosimeters lots together with an analysis of the shine curves and of calibration selecting those that presented a better behaviour and a more rea liable response. (Author)

  15. Comparison between two kind of power circuits for personal dosimeter

    International Nuclear Information System (INIS)

    Liu Zhengshan; Deng Changming; Guo Zhanjie

    2002-01-01

    Personal Dosimeter is commonly requested using battery for its power supply, and hope the battery life is long. Also with the fall of battery voltage, some performance of instrument as well as drop, Reasonable supply design can protract the battery life. The author introduces two method: power supply with battery directly and supply used power chip conversion. Combine personal dosimeter, the authors carried comparison for battery life, power consumption, cost and volume. Based on the comparison result and instrument fact request, one can choose method of power circuit

  16. measurement of the supralinearity of 7LiF thermoluminescent dosimeters

    International Nuclear Information System (INIS)

    Hancock, I.B.

    1978-04-01

    A study has been made of the supralinearity of the 7 LiF thermoluminesecent dosimeter (T.L.D.) used in the UKAEA personnel criticality dosimeter by exposure to a 60 Co source. the response of the T.L.D. was linear up to approximately 450 rad, above which it became increasingly supralinear, over-reading by about 50% at 2000 rad. The results have been used to produce a correction graph, as a function of apparent dose up to 3500 rad, for routine use in nuclear accident dosimetry. (author)

  17. Intercomparison of radiation dosimeters for individual monitoring

    International Nuclear Information System (INIS)

    1993-05-01

    The Co-ordinated Research Programme on Intercomparison for Individual Monitoring was established to provide participants with an opportunity to assess (1) their ability to measure external photon radiation fields and (2) the potential impact of introduction of the new operational quantities on their dosimetry programmes. Twenty-four laboratories from 18 IAEA Member States and three international organizations, including the IAEA, participated. The results of phase II of the CRP are presented in this document, which includes a compilation of the presentations and conclusions from the meeting. Refs, figs and tabs

  18. Measurement of the dose by dispersed radiation in a lineal accelerator using thermoluminescent dosimeters of CaSO4:Dy

    International Nuclear Information System (INIS)

    Chavez C, N.; Torijano, E.; Azorin, J.; Herrera, A.

    2014-08-01

    The thermoluminescence (Tl) is based on the principle of the luminescent in a material when is heated below their incandescence temperature. Is a technique very used in dosimetry that is based on the property that have most of the crystalline materials regarding the storage of the energy that they absorb when are exposed to the ionizing radiations. When this material has been irradiated previously, the radioactive energy that contains is liberated in form of light. In general, the principles that govern the thermoluminescence are in essence the same of those responsible for all the luminescent processes and, this way, the thermoluminescence is one of the processes that are part of the luminescence phenomenon. For this work, the dispersed radiation was measured in the therapy area of the lineal accelerator of medical use type Elekta, using thermoluminescent dosimeters of CaSO 4 :Dy + Ptfe developed and elaborated in the Universidad Autonoma Metropolitana, Unidad Iztapalapa. With the dosimeters already characterized and calibrated, we proceeded to measure the dispersed radiation being a patient in treatment. The results showed values for the dispersed radiation the order of a third of the dose received by the patient on the treatment table at 30 cm of the direct beam and the order of a hundredth in the control area (4 m of the direct beam, approximately). The conclusion is that the thermoluminescent dosimeters of CaSO 4 : Dy + Ptfe are appropriate to measure dispersed radiation dose in radiotherapy. (author)

  19. TLD system for the monitoring of the environmental radioactivity

    International Nuclear Information System (INIS)

    Stochioiu, Ana; Sahagia, Maria; Tudor, Ion

    2008-01-01

    The paper presents a high sensitivity TLD system, designed for the survey of the environmental radioactivity. It is based on the use of TL detectors type LiF:Mg, Cu, P, commercially known as GR-200A. The dosimeter designed in our Institute, contains 3 detectors, and the measurement value is calculated as the arithmetic mean. A very sensitive, TL Reader, READER ANALYSER RA'94 was chosen and an optimal thermal cycle was designed, such as to enhance the measurement performances. For each placement, a set of 3 dosemeters is used, and survey intervals from 1 to 100 days, depending on the radioactivity level and reporting requirements, are selected. The technical characteristics of the system were determined by exposing the dosimeters in reference X and gamma radiation fields, such as required by the IEC standard 61066:iun.2006 'Thermoluminescence dosimetry systems for personal and environmental monitoring'. The main technical parameters are of highest quality and recommend it for use in the survey of the environmental radioactivity, at the level of ambient dose equivalent rate, due to normal natural radioactivity, in open areas. The paper describes the method of characterisation and measurement results, as well as their relevance. (author)

  20. Measurement of the external doses at low and high exposures by Agfa personal monitoring film, FD-III-B badge

    International Nuclear Information System (INIS)

    Mihai, F.; Stochioiu, A.; Bercea, S.; Udup, E.; Tudor, I.

    2008-01-01

    Full text: A growing number of papers report occupational exposure monitoring by different dosemeter devices. Researches into improvement of the personal dosemeter systems are encouraged. Dose measurement and especially skin dose measurement, are therefore increasingly important. Methods and acceptable dosemeter are not clearly defined and differ from a dosimetry laboratory to other depending on their specifically work procedure and type of dosemeter utilized. Taking into account the characteristic of the work place (industry, research, medical application) dosemeter have to cover a large dose range. The dose assessment problems are manly in low doses range (under 0.2 mSv) and higher doses range. For example, typical for radiation oncology and nuclear accident dosimetry the dose ranges are on the order of 2-70 Gy and 0.1-5 Gy. This work presents the performance of the Agfa personal monitoring film, badge FD-III-B type used in the most of the occupational exposure monitoring from Romania. The FD-III-B dosimeter, manufactured in Romania, is used for monitoring personnel working in radioactive environment - the photodosimetric method is the one of the most reliable method presenting the advantage that the film may be considered as master probe over 30 year and may be utilized during litigation. The dosimeter contains metallic filters of Al, Cu, Pb with different thicknesses, a plastic filter and a open window. The badge FD-III-B is a class B dosimeter according to the dose range and an indices 3 dosimeter from energy range of the radiations in conformity with Romanian Standard in force. Film Agfa 'personal monitoring', tacked in this experiment, consists of a low speed film (D2) and a very sensitive film (D10) designed for recording X-ray, gamma and beta radiation over the doses range 0.1 mSv to 1 Sv. The experiments have been effectuated at different background optical density (base fog) of the films and at different sources of radiation: 13 :7Cs, 60 Co and 24 : 1 Am