WorldWideScience

Sample records for accredited dosimetry calibration

  1. QA experience at the University of Wisconsin accredited dosimetry calibration laboratory

    Energy Technology Data Exchange (ETDEWEB)

    DeWard, L.A.; Micka, J.A. [Univ. of Wisconsin, Madison, WI (United States)

    1993-12-31

    The University of Wisconsin Accredited Dosimetry Calibration Laboratory (UW ADCL) employs procedure manuals as part of its Quality Assurance (QA) program. One of these manuals covers the QA procedures and results for all of the UW ADCL measurement equipment. The QA procedures are divided into two main areas: QA for laboratory equipment and QA for external chambers sent for calibration. All internal laboratory equipment is checked and recalibrated on an annual basis, after establishing its consistency on a 6-month basis. QA for external instruments involves checking past calibration history as well as comparing to a range of calibration values for specific instrument models. Generally, the authors find that a chamber will have a variation of less than 0.5 % from previous Co-60 calibration factors, and falls within two standard deviations of previous calibrations. If x-ray calibrations are also performed, the energy response of the chamber is plotted and compared to previous instruments of the same model. These procedures give the authors confidence in the transfer of calibration values from National Institute of Standards and Technology (NIST).

  2. History, organization, and oversight of the accredited dosimetry calibration laboratories by the AAPM

    Energy Technology Data Exchange (ETDEWEB)

    Rozenfeld, M. [St. James Hospital and Health Centers, Chicago Heights, IL (United States)

    1993-12-31

    For more than 20 years, the American Association of Physicists in Medicine (AAPM) has operated an accreditation program for secondary standards laboratories that calibrate radiation measuring instruments. Except for one short period, that program has been able to provide the facilities to satisfy the national need for accurate calibrations of such instruments. That exception, in 1981, due to the combination of the U.S. Nuclear Regulatory Commission (NRC) requiring instrument calibrations by users of cobalt-60 teletherapy units and the withdrawal of one of the three laboratories accredited at that time. However, after successful operation as a Task Group of the Radiation Therapy Committee (RTC) of the AAPM for two decades, a reorganization of this structure is now under serious consideration by the administration of the AAPM.

  3. Dosimetry and Calibration Section

    International Nuclear Information System (INIS)

    The two tasks of the Dosimetry and Calibration Section at CERN are the Individual Dosimetry Service which assures the personal monitoring of about 5000 persons potentially exposed to ionizing radiation at CERN, and the Calibration Laboratory which verifies all the instruments and monitors. This equipment is used by the sections of the RP Group for assuring radiation protection around CERN's accelerators, and by the Environmental Section of TISTE. In addition, nearly 250 electronic and 300 quartz fibre dosimeters, employed in operational dosimetry, are calibrated at least once a year. The Individual Dosimetry Service uses an extended database (INDOS) which contains information about all the individual doses ever received at CERN. For most of 1997 it was operated without the support of a database administrator as the technician who had assured this work retired. The Software Support Section of TIS-TE took over the technical responsibility of the database, but in view of the many other tasks of this Section and the lack of personnel, only a few interventions for solving immediate problems were possible

  4. Individual dosimetry and calibration

    International Nuclear Information System (INIS)

    In 1995 both the Individual Dosimetry and Calibration Sections worked under the condition of a status quo and concentrated fully on the routine part of their work. Nevertheless, the machine for printing the bar code which will be glued onto the film holder and hence identify the people when entering into high radiation areas was put into operation and most of the holders were equipped with the new identification. As far as the Calibration Section is concerned the project of the new source control system that is realized by the Technical Support Section was somewhat accelerated

  5. HPS instrument calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Masse, F.X; Eisenhower, E.H.; Swinth, K.L.

    1993-12-31

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory.

  6. HPS instrument calibration laboratory accreditation program

    International Nuclear Information System (INIS)

    The purpose of this paper is to provide an accurate overview of the development and structure of the program established by the Health Physics Society (HPS) for accrediting instrument calibration laboratories relative to their ability to accurately calibrate portable health physics instrumentation. The purpose of the program is to provide radiation protection professionals more meaningful direct and indirect access to the National Institute of Standards and Technology (NIST) national standards, thus introducing a means for improving the uniformity, accuracy, and quality of ionizing radiation field measurements. The process is designed to recognize and document the continuing capability of each accredited laboratory to accurately perform instrument calibration. There is no intent to monitor the laboratory to the extent that each calibration can be guaranteed by the program; this responsibility rests solely with the accredited laboratory

  7. Accreditation ISO/IEC 1705 in dosimetry: Experience and results

    International Nuclear Information System (INIS)

    The objective of this work is to present the experience in the process of accreditation of the radiation dosimetry service in which there are trials for the determination of radiation doses due to internal and external exhibitions. Is They describe the aspects that were considered for the design and development of a system of quality and results after its implementation. A review of the benefits accreditation has been reported to the organization is finally made. (Author)

  8. CIEMAT EXTERNAL DOSIMETRY SERVICE: ISO/IEC 17025 ACCREDITATION AND 3 Y OF OPERATIONAL EXPERIENCE AS AN ACCREDITED LABORATORY.

    Science.gov (United States)

    Romero, A M; Rodríguez, R; López, J L; Martín, R; Benavente, J F

    2016-09-01

    In 2008, the CIEMAT Radiation Dosimetry Service decided to implement a quality management system, in accordance with established requirements, in order to achieve ISO/IEC 17025 accreditation. Although the Service comprises the approved individual monitoring services of both external and internal radiation, this paper is specific to the actions taken by the External Dosimetry Service, including personal and environmental dosimetry laboratories, to gain accreditation and the reflections of 3 y of operational experience as an accredited laboratory. PMID:26567323

  9. The method validation step of biological dosimetry accreditation process

    Energy Technology Data Exchange (ETDEWEB)

    Roy, L.; Voisin, P.A.; Guillou, A.C.; Busset, A.; Gregoire, E.; Buard, V.; Delbos, M.; Voisin, Ph. [Institut de Radioprotection et de Surete Nucleaire, LDB, 92 - Fontenay aux Roses (France)

    2006-07-01

    One of the missions of the Laboratory of Biological Dosimetry (L.D.B.) of the Institute for Radiation and Nuclear Safety (I.R.S.N.) is to assess the radiological dose after an accidental overexposure suspicion to ionising radiation, by using radio-induced changes of some biological parameters. The 'gold standard' is the yield of dicentrics observed in patients lymphocytes, and this yield is converted in dose using dose effect relationships. This method is complementary to clinical and physical dosimetry, for medical team in charge of the patients. To obtain a formal recognition of its operational activity, the laboratory decided three years ago, to require an accreditation, by following the recommendations of both 17025 General Requirements for the Competence of Testing and Calibration Laboratories and 19238 Performance criteria for service laboratories performing biological dosimetry by cyto-genetics. Diagnostics, risks analysis were realized to control the whole analysis process leading to documents writing. Purchases, personnel department, vocational training were also included in the quality system. Audits were very helpful to improve the quality system. One specificity of this technique is that it is not normalized therefore apart from quality management aspects, several technical points needed some validations. An inventory of potentially influent factors was carried out. To estimate their real effect on the yield of dicentrics, a Placket-Burman experimental design was conducted. The effect of seven parameters was tested: the BUdr (bromodeoxyuridine), PHA (phytohemagglutinin) and colcemid concentration, the culture duration, the incubator temperature, the blood volume and the medium volume. The chosen values were calculated according to the uncertainties on the way they were measured i.e. pipettes, thermometers, test tubes. None of the factors has a significant impact on the yield of dicentrics. Therefore the uncertainty linked to their use was

  10. Requirements for the accreditation of a calibration laboratory

    International Nuclear Information System (INIS)

    CNEA's activity in calibration is recent but it has a significant development. To assure high quality results, activity must be sustained and improved from day to day. The calibrations laboratory was accredited before Laboratories Qualification Committee, thus adding reliability to its results and making it more competitive when compared to other laboratories not accredited. Among other services given are supervision and follow up of calibrations in laboratories, participation in interlaboratory assays together with other calibration laboratories and assessments on calibration aspects of measuring equipment. (author)

  11. US Department of Energy Laboratory Accredition Program (DOELAP) for personnel dosimetry systems

    Energy Technology Data Exchange (ETDEWEB)

    Cummings, F.M.; Carlson, R.D.; Loesch, R.M.

    1993-12-31

    Accreditation of personnel dosimetry systems is required for laboratories that conduct personnel dosimetry for the U.S. Department of Energy (DOE). Accreditation is a two-step process which requires the participant to pass a proficiency test and an onsite assessment. The DOE Laboratory Accreditation Program (DOELAP) is a measurement quality assurance program for DOE laboratories. Currently, the DOELAP addresses only dosimetry systems used to assess the whole body dose to personnel. A pilot extremity DOELAP has been completed and routine testing is expected to begin in January 1994. It is expected that participation in the extremity program will be a regulatory requirement by January 1996.

  12. Calibration methods of plane-parallel ionization chambers used in electron dosimetry

    International Nuclear Information System (INIS)

    The use of linear accelerators in radiotherapy is of great importance in Medicine, and according to international recommendations the electron beam dosimetry has to be performed using plane-parallel ionization chambers, previously calibrated in standard gamma radiation fields at accredited laboratories. In this work, calibration methods of plane-parallel ionization chambers used in dosimetry procedures of high energy electron beams of clinical accelerators were presented, tested and intercompared. The experiments were carried out using gamma radiation beams of 60 Co at the Calibration Laboratory of Clinical Dosemeters at IPEN and electron beams od 4 to 16 MeV at the Radiotherapy Department of Hospital Israelita Albert Einstein, Sao Paulo. A method was chosen to be established at IPEN. Proposals of the calibration procedure, calibration certificate and data sheets are presented. (author)

  13. High-dose secondary calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Humphreys, J.C. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program.

  14. Dosimetry intercomparisons in European medical device sterilization plants

    DEFF Research Database (Denmark)

    Miller, A.; Sharpe, P.H.G.

    2000-01-01

    Dosimetry intercomparisons have been carried out involving two-thirds of all European radiation sterilization facilities. Dosimeters for the intercomparisons were supplied by two accredited calibration laboratories. The results show good agreement, and indicate overall dosimetry accuracy of the o...

  15. Accreditation of a personal dosimetry service in Switzerland: practical experience and transition from EN 45004 to ISO 17025.

    Science.gov (United States)

    Boschung, M; Wernli, C

    2001-01-01

    In compliance with the Swiss legislation on radiological protection, the Paul Scherrer Institute (PSI) operates a dosimetry service that is approved by the Swiss Federal Nuclear Safety Inspectorate. In 1997, the dosimetry service was also accredited by the Swiss Federal Office of Metrology and Accreditation as an inspection body for legal personal and environmental dosimetry, according to EN 45004. The accreditation covers determination of personal dose equivalent for photon, neutron and beta radiation, and ambient dose equivalent for photon and neutron radiation, by means of thermoluminescence and solid state track detection techniques. Within this formal accreditation it was confirmed that the relevant requirements of ISO 9002 are also fulfilled. The first re-accreditation will take place in 2001 and work is going on to achieve the transition from EN 45004 to ISO 17025. Accreditation is a feasible, practicable and acceptable way to achieve harmonisation in the field of dosimetry. However, before starting on the path to formal accreditation, a careful analysis should be made, taking into consideration not only cost-benefit aspects but also national legal requirements. PMID:11586715

  16. Accreditation of a personal dosimetry service in Switzerland: Practical experience and transition from EN 45004 to ISO 17025

    Energy Technology Data Exchange (ETDEWEB)

    Boschung, M.; Wernli, C

    2001-07-01

    In compliance with the Swiss legislation on radiological protection, the Paul Scherrer Institute (PSI) operates a dosimetry service that is approved by the Swiss Federal Nuclear Safety Inspectorate. In 1997, the dosimetry service was also accredited by the Swiss Federal Office of Metrology and Accreditation as an inspection body for legal personal and environmental dosimetry, according to EN 45004. The accreditation covers determination of personal dose equivalent for photon, neutron and beta radiation, and ambient dose equivalent for photon and neutron radiation, by means of thermoluminescence and solid state track detection techniques. Within this formal accreditation it was confirmed that the relevant requirements of ISO 9002 are also fulfilled. The first re-accreditation will take place in 2001 and work is going on to achieve the transition from EN 45004 to ISO 17025. Accreditation is a feasible, practicable and acceptable way to achieve harmonisation in the field of dosimetry. However, before starting on the path to formal accreditation, a careful analysis should be made, taking into consideration not only cost-benefit aspects but also national legal requirements. (author)

  17. Accreditation ISO/IEC 1705 in dosimetry: Experience and results; Acreditacion ISO/IEC 17025 en dosimetria: Experiencia y resultados

    Energy Technology Data Exchange (ETDEWEB)

    Martin Garcia, R.; Navarro Bravo, T.

    2013-07-01

    The objective of this work is to present the experience in the process of accreditation of the radiation dosimetry service in which there are trials for the determination of radiation doses due to internal and external exhibitions. Is They describe the aspects that were considered for the design and development of a system of quality and results after its implementation. A review of the benefits accreditation has been reported to the organization is finally made. (Author)

  18. Establishment of qualities mammography according to the standard IEC-61267 in the laboratory of metrology of ionizing radiation of the National Center of Dosimetry and enlargement of the accreditation by ENAC; Establecimiento de las calidades de mamografia segun la norma IEC-61267 en el laboratorio de metrologia de radiaciones ionizantes del centro nacional de dosimetria (CND) y proceso de ampliacion de su acreditacion por ENAC

    Energy Technology Data Exchange (ETDEWEB)

    Roig Petit, F.; Mestre de Juan, V.; Alabau Albors, J.; Palma Copete, J.; Ruiz Rodriguez, J. C.; Pons Mocholi, S.

    2013-07-01

    The extension of the accreditation of the laboratory of the National Center Dosimetry (No. 58/LC10.036) by the national accreditation entity (ENAC), according to the ISO 17025 standard [2], for the standard qualities of mammography by the IEC 61267 comes to meet part of the needs that demand our health care environment in terms of radiation measuring instruments calibration. This work intends to publicize this enlargement commenting on the different phases of the process to get the accreditation. (Author)

  19. Accreditation of the Personal Dosimetry internal Service Tecnatom by the National Entity (ENAC); Acreditacion del Servicio de Dosimetria Personal Interna de Tecnatom por la Entidad Nacional de Acreditacion (ENAC)

    Energy Technology Data Exchange (ETDEWEB)

    Bravo, B.; Marchena, P.

    2014-07-01

    The service of personal Dosimetry internal Tecnatom has made the process of adapting its methodology and quality assurance, requirements technical and management will be required to obtain accreditation from the National Accreditation Entity according to ISO / IEC 170251 standard {sup G}eneral Requirements competence of testing and calibration laboratories. To carry out this process, the laboratory has defined quality criteria set out in their test procedures, based on ISO Standards 27048: 2011; ISO 20553: 2005 and ISO 28218: 2010. This paper describes what has been the methodology used to implement the requirements of different ISO test methods of SDPI Tecnatom. (Author)

  20. Reliability, calibration and metrology in ionizing radiation dosimetry

    CERN Document Server

    Cundin, Luisiana X

    2013-01-01

    Radiation dosimetry systems are complex systems, comprised of a milieu of components, designed for determining absorbed dose after exposure to ionizing radiation. Although many materials serve as absorbing media for measurement, thermoluminescent dosimeters represent some of the more desirable materials available; yet, reliability studies have revealed a clear and definite decrement in dosimeter sensitivity after repeated use. Unfortunately, repeated use of any such material for absorbing media in ionizing radiation dosimetry will in time experience performance decrements; thus, in order to achieve the most accuracy and/or precision in dosimetry, it is imperative proper compensation be made in calibration. Yet, analysis proves the majority of the measured decrement in sensitivity experienced by dosimeters is attributable to drift noise and not to any degradation in dosimeter performance, at least, not to any great degree. In addition to investigating dosimeter reliability, implications for metrological tracea...

  1. Dosimetry through the Secondary Laboratory of Dosimetric Calibration of Mexico

    International Nuclear Information System (INIS)

    In the beginnings of the sixty years an urgent necessity is presented mainly in the developing countries, of improving in important form the accuracy in the dosimetry of external faces in therapy of radiations (radiotherapy centers), mainly in the calibration of clinical dosemeters. In 1976 the International Atomic Energy Agency, (IAEA), and the World Health Organization, (WHO), they carried out a mutual agreement with regard to the establishment and operation of a net of Secondary Patron Laboratories of Dosimetry, (LSCD). The necessity to establish measure patterns in the field of the dosimetry of the ionizing radiations, is necessary, to have an accuracy but high in the dosimetry of the radiation beams in therapy which is highly dependent of the dose given to the tumor of those patient with cancer. Similar levels of accuracy are required in protection measures to the radiation with an acceptable smaller accuracy, however, when the personal dosemeters are used to determine the doses received by the individuals under work conditions, such mensurations in therapy of radiations and radiological protection will have traceability through a chain of comparisons to primary or national patterns. The traceability is necessary to assure the accuracy and acceptability of the dosimetric measures, as well as, the legal and economic implications. The traceability is also necessary in the dosimetry of high dose like in the sterilization of different products. The main function of the LSCD is to provide a service in metrology of ionizing radiations, maintaining the secondary or national patterns, which have a traceability to the International System of measures, which is based for if same in the comparison of patterns in the Primary Laboratories of Dosimetry (LPD) under the auspice of the International Office of Weights and Measure (BIPM). The secondary and national patterns in the LSCD constitute in Mexico, the national patterns of the magnitudes in the dosimetry of the ionizing

  2. Secondary calibration laboratory for ionizing radiation laboratory accreitation program National Institute of Standards and Technology National Voluntary Laboratory Accreditation Program

    Energy Technology Data Exchange (ETDEWEB)

    Martin, P.R.

    1993-12-31

    This paper presents an overview of the procedures and requirements for accreditation under the Secondary Calibration Laboratory for Ionizing Radiation Program (SCLIR LAP). The requirements for a quality system, proficiency testing and the onsite assessment are discussed. The purpose of the accreditation program is to establish a network of secondary calibration laboratories that can provide calibrations traceable to the primary national standards.

  3. Calibration methods of plane-parallel ionization chambers used in electron dosimetry; Metodos de calibracao de camaras de ionizacao de placas paralelas para dosimetria de feixes de eletrons

    Energy Technology Data Exchange (ETDEWEB)

    Bulla, Roseli Tadeu

    1999-07-01

    The use of linear accelerators in radiotherapy is of great importance in Medicine, and according to international recommendations the electron beam dosimetry has to be performed using plane-parallel ionization chambers, previously calibrated in standard gamma radiation fields at accredited laboratories. In this work, calibration methods of plane-parallel ionization chambers used in dosimetry procedures of high energy electron beams of clinical accelerators were presented, tested and intercompared. The experiments were carried out using gamma radiation beams of {sup 60} Co at the Calibration Laboratory of Clinical Dosemeters at IPEN and electron beams od 4 to 16 MeV at the Radiotherapy Department of Hospital Israelita Albert Einstein, Sao Paulo. A method was chosen to be established at IPEN. Proposals of the calibration procedure, calibration certificate and data sheets are presented. (author)

  4. Accreditation of a system of extremity dosimetry: validation and uncertainty of method; Acreditacion de un sistema de dosimetria de extremidades: validacion e incertidumbre del metodo

    Energy Technology Data Exchange (ETDEWEB)

    Romero Gutierrez, A. M.; Rodriguez Jimenez, R.; Lopez Moyano, J. L.

    2013-07-01

    The authors' goal is to spread the practical experience gained during the accreditation process paying special attention to the process of method validation and estimation uncertainty of the dosimetry system. (Author)

  5. Generalized EPID calibration for in vivo transit dosimetry.

    Science.gov (United States)

    Fidanzio, Andrea; Cilla, Savino; Greco, Francesca; Gargiulo, Laura; Azario, Luigi; Sabatino, Domenico; Piermattei, Angelo

    2011-01-01

    Many researchers are studying new in vivo dosimetry methods based on the use of Elelctronic portal imaging devices (EPIDs) that are simple and efficient in their daily use. However the need of time consuming implementation measurements with solid water phantoms for the in vivo dosimetry implementation can discourage someone in their use. In this paper a procedure has been proposed to calibrate aSi EPIDs for in vivo transit dosimetry. The dosimetric equivalence of three aSi Varian EPIDs has been investigated in terms of signal reproducibility and long term stability, signal linearity with MU and dose per pulse and signal dependence on the field dimensions. The signal reproducibility was within ± 0.5% (2SD), while the long term signal stability has been maintained well within ± 2%. The signal linearity with the monitor units (MU) was within ± 2% and within ± 0.5% for the EPIDs controlled by the IAS 2, and IAS 3 respectively. In particular it was verified that the correction factor for the signal linearity with the monitor units, k(lin), is independent of the beam quality, and the dose per pulse absorbed by the EPID. For 6, 10 and 15 MV photon beams, a generalized set of correlation functions F(TPR,w,L) and empirical factors f(TPR,d,L) as a function of the Tissue Phantom Ratio (TPR), the phantom thickness, w, the square field side, L, and the distance, d, between the phantom mid-plane and the isocentre were determined to reconstruct the isocenter dose. The tolerance levels of the present in vivo dosimetry method ranged between ± 5% and ± 6% depending on the tumor body location. In conclusion, the procedure proposed, that use generalized correlation functions, reduces the effort for the in vivo dosimetry method implementation for those photon beams with TPR within ± 0.3% as respect those here used.

  6. Panoptes: Calibration of a dosimetry system for eye brachytherapy

    International Nuclear Information System (INIS)

    Intraocular cancer is a serious threat to the lives of those that suffer from it. Dosimetry for eye brachytherapy presents a significant challenge due to the inherently steep dose gradients that are needed to treat such small tumours in close proximity to sensitive normal structures. This issue is addressed by providing much needed quality assurance to eye brachytherapy, a novel volumetric dosimetry system, called PANOPTES was developed. This study focuses on the preliminary characterisation and calibration of the system. Using ion beam facilities, the custom, pixelated silicon detector of PANOPTES was shown to have good charge collection uniformity and a well defined sensitive volume. Flat-field calibration was conducted on the device using a 250 kVp orthovoltage beam. Finally, the detector and phantom were simulated with Monte Carlo in Geant4, to create water equivalent dose correction factors for each pixel across a range of angles. - Highlights: • Volumetric detector system produced for plaque brachytherapy. • Orthovoltage, flat-field calibration performed for detector pixels. • Monte Carlo simulation showed mostly little angular deviation across all angles. • Ion beam induced charge collection showed pixels uniform and fully depleted

  7. The U.S. Department of Energy Laboratory Accreditation Program for testing the performance of extremity dosimetry systems: a summary of the program status

    International Nuclear Information System (INIS)

    In 1986, The U.S. Department of Energy (DOE) implemented a program to test the performance of its personnel whole-body dosimetry systems. This program was the DOE Laboratory Accreditation Program (DOELAP). The program parallels the performance testing program specified in the American National Standard for Dosimetry - Personnel Dosimetry Performance -Criteria for Testing (ANSI N13.11-1983), but also addresses the additional dosimetry needs of DOE facilities. As an extension of the whole-body performance testing program, the DOE is now developing a program to test the performance of personnel extremity dosimetry systems. The draft DOE standard for testing extremity dosimetry systems is much less complex than the whole-body dosimetry standard and reflects the limitations imposed on extremity dosimetry by dosimeter design and irradiation geometry. A pilot performance test session has been conducted to evaluate the proposed performance-testing standard. (author)

  8. Guidelines for the Calibration of Routine Dosimetry Systems for use in Radiation Processing

    DEFF Research Database (Denmark)

    Sharpe, Peter; Miller, Arne

    A set of guidelines has been developed to assist in the calibration of routine dosimetry systems for use in industrial radiation processing plants. Topics covered include the calibration of equipment, the performance of calibration irradiations and the derivation of mathematical functions...... to represent the calibration. Guidance is also given on methods for the estimation of uncertainty....

  9. Comparison of pencil-type ionization chamber calibration results and methods between dosimetry laboratories.

    Science.gov (United States)

    Hourdakis, Costas J; Büermann, Ludwig; Ciraj-Bjelac, Olivera; Csete, Istvan; Delis, Harry; Gomola, Igor; Persson, Linda; Novak, Leos; Petkov, Ivailo; Toroi, Paula

    2016-01-01

    A comparison of calibration results and procedures in terms of air kerma length product, PKL, and air kerma, K, was conducted between eight dosimetry laboratories. A pencil-type ionization chamber (IC), generally used for computed tomography dose measurements, was calibrated according to three calibration methods, while its residual signal and other characteristics (sensitivity profile, active length) were assessed. The results showed that the "partial irradiation method" is the preferred method for the pencil-type IC calibration in terms of PKL and it could be applied by the calibration laboratories successfully. Most of the participating laboratories achieved high level of agreement (>99%) for both dosimetry quantities (PKL and K). Estimated relative standard uncertainties of comparison results vary among laboratories from 0.34% to 2.32% depending on the quantity, beam quality and calibration method applied. Detailed analysis of the assigned uncertainties have been presented and discussed.

  10. Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry

    International Nuclear Information System (INIS)

    Purpose: The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by ∼25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. Methods: The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Results: Calibration precision was measured to be better than 5%–7%, 3%–5%, and 2%–4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.7 ± 1.1 mV cGy−1 versus the CT scatter phantom 29.2 ± 1.0 mV cGy−1 and FIA with x-ray 29.9 ± 1.1 mV cGy−1 methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of ∼3000 mV. Conclusions: The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration coefficients for the eventual

  11. Establishing a standard calibration methodology for MOSFET detectors in computed tomography dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Brady, S. L.; Kaufman, R. A. [Department of Radiological Sciences, St. Jude Children' s Research Hospital, Memphis, Tennessee 38105 (United States)

    2012-06-15

    Purpose: The use of metal-oxide-semiconductor field-effect transistor (MOSFET) detectors for patient dosimetry has increased by {approx}25% since 2005. Despite this increase, no standard calibration methodology has been identified nor calibration uncertainty quantified for the use of MOSFET dosimetry in CT. This work compares three MOSFET calibration methodologies proposed in the literature, and additionally investigates questions relating to optimal time for signal equilibration and exposure levels for maximum calibration precision. Methods: The calibration methodologies tested were (1) free in-air (FIA) with radiographic x-ray tube, (2) FIA with stationary CT x-ray tube, and (3) within scatter phantom with rotational CT x-ray tube. Each calibration was performed at absorbed dose levels of 10, 23, and 35 mGy. Times of 0 min or 5 min were investigated for signal equilibration before or after signal read out. Results: Calibration precision was measured to be better than 5%-7%, 3%-5%, and 2%-4% for the 10, 23, and 35 mGy respective dose levels, and independent of calibration methodology. No correlation was demonstrated for precision and signal equilibration time when allowing 5 min before or after signal read out. Differences in average calibration coefficients were demonstrated between the FIA with CT calibration methodology 26.7 {+-} 1.1 mV cGy{sup -1} versus the CT scatter phantom 29.2 {+-} 1.0 mV cGy{sup -1} and FIA with x-ray 29.9 {+-} 1.1 mV cGy{sup -1} methodologies. A decrease in MOSFET sensitivity was seen at an average change in read out voltage of {approx}3000 mV. Conclusions: The best measured calibration precision was obtained by exposing the MOSFET detectors to 23 mGy. No signal equilibration time is necessary to improve calibration precision. A significant difference between calibration outcomes was demonstrated for FIA with CT compared to the other two methodologies. If the FIA with a CT calibration methodology was used to create calibration

  12. Calibration of a tertiary standard in N-ISO qualities for radioprotection and personal dosimetry

    International Nuclear Information System (INIS)

    Dosimetric calibration of radiation monitors and personal dosimeters in different radiological quantities are performed in order to obtain accurate measurements, for this reason the SSDL calculates the dosimetry calibration factor and its associated uncertainty, for each range of use. The calibration factor is performed using the known radiation field method and its uncertainty is calculated according to the ISO recommendations. The SSDL calculates the expanded uncertainty (Uc) with a coverage factor that provides a level of not less than 95 % of confidence. (authors).

  13. Validating dose rate calibration of radiotherapy photon beams through IAEA/WHO postal audit dosimetry service

    International Nuclear Information System (INIS)

    In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World Health Organisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the low-energy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility. (author)

  14. Calibration of a He accumulation fluence monitor for fast reactor dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Ito, Chikara [Power Reactor and Nuclear Fuel Development Corp., Oarai, Ibaraki (Japan). Oarai Engineering Center

    1997-03-01

    The helium accumulation fluence monitor (HAFM) has been developed for a fast reactor dosimetry. The HAFM measurement system was calibrated using He gas and He implanted samples and the measurement accuracy was confirmed to be less than 5%. Based on the preliminary irradiation test in JOYO, the measured He in the {sup 10}B type HAFM agreed well with the calculated values using the JENDL-3.2 library. (author)

  15. A round-robin gamma stereotactic radiosurgery dosimetry interinstitution comparison of calibration protocols

    Energy Technology Data Exchange (ETDEWEB)

    Drzymala, R. E., E-mail: drzymala@wustl.edu [Department of Radiation Oncology, Washington University, St. Louis, Missouri 63110 (United States); Alvarez, P. E. [Imaging and Radiation Oncology Core Houston, UT MD Anderson Cancer Center, Houston, Texas 77030 (United States); Bednarz, G. [Radiation Oncology Department, University of Pittsburgh Medical Center, Pittsburgh, Pennsylvania 15232 (United States); Bourland, J. D. [Department of Radiation Oncology, Wake Forest University, Winston-Salem, North Carolina 27157 (United States); DeWerd, L. A. [Department of Medical Physics, University of Wisconsin-Madison, Madison, Wisconsin 53705 (United States); Ma, L. [Department of Radiation Oncology, University California San Francisco, San Francisco, California 94143 (United States); Meltsner, S. G. [Department of Radiation Oncology, Duke University Medical Center, Durham, North Carolina 27710 (United States); Neyman, G. [Department of Radiation Oncology, The Cleveland Clinic Foundation, Cleveland, Ohio 44195 (United States); Novotny, J. [Medical Physics Department, Hospital Na Homolce, Prague 15030 (Czech Republic); Petti, P. L. [Gamma Knife Center, Washington Hospital Healthcare System, Fremont, California 94538 (United States); Rivard, M. J. [Department of Radiation Oncology, Tufts University School of Medicine, Boston, Massachusetts 02111 (United States); Shiu, A. S. [Department of Radiation Oncology, University of Southern California, Los Angeles, California 90033 (United States); Goetsch, S. J. [San Diego Medical Physics, Inc., La Jolla, California 92037 (United States)

    2015-11-15

    Purpose: Absorbed dose calibration for gamma stereotactic radiosurgery is challenging due to the unique geometric conditions, dosimetry characteristics, and nonstandard field size of these devices. Members of the American Association of Physicists in Medicine (AAPM) Task Group 178 on Gamma Stereotactic Radiosurgery Dosimetry and Quality Assurance have participated in a round-robin exchange of calibrated measurement instrumentation and phantoms exploring two approved and two proposed calibration protocols or formalisms on ten gamma radiosurgery units. The objectives of this study were to benchmark and compare new formalisms to existing calibration methods, while maintaining traceability to U.S. primary dosimetry calibration laboratory standards. Methods: Nine institutions made measurements using ten gamma stereotactic radiosurgery units in three different 160 mm diameter spherical phantoms [acrylonitrile butadiene styrene (ABS) plastic, Solid Water, and liquid water] and in air using a positioning jig. Two calibrated miniature ionization chambers and one calibrated electrometer were circulated for all measurements. Reference dose-rates at the phantom center were determined using the well-established AAPM TG-21 or TG-51 dose calibration protocols and using two proposed dose calibration protocols/formalisms: an in-air protocol and a formalism proposed by the International Atomic Energy Agency (IAEA) working group for small and nonstandard radiation fields. Each institution’s results were normalized to the dose-rate determined at that institution using the TG-21 protocol in the ABS phantom. Results: Percentages of dose-rates within 1.5% of the reference dose-rate (TG-21 + ABS phantom) for the eight chamber-protocol-phantom combinations were the following: 88% for TG-21, 70% for TG-51, 93% for the new IAEA nonstandard-field formalism, and 65% for the new in-air protocol. Averages and standard deviations for dose-rates over all measurements relative to the TG-21 + ABS

  16. Establishment of a new calibration method of pencil ionization chamber for dosimetry in computed tomography

    International Nuclear Information System (INIS)

    Pencil ionization chambers are used for beam dosimetry in computed tomography equipment (CT). In this study, a new calibration methodology was established, in order to make the Calibration Laboratory of Instituto de Pesquisas Energeticas e Nucleares (LCI) suitable to international metrological standards, dealing with specific procedures for calibration of these chambers used in CT. Firstly, the setup for the new RQT radiation qualities was mounted, in agreement with IEC61267 from the International Electrotechnical Commission (IEC). After the establishment of these radiation qualities, a specific calibration methodology for pencil ionization chambers was set, according to Technical Report Series No. 457, from the International Atomic Energy Agency (IAEA), which describes particularities of the procedure to be followed by the Secondary Standard Dosimetry Laboratories (SSDL's), concerning to collimation and positioning related to the radiation beam. Initially, PPV (kV) measurements and the determination of copper additional filtrations were carried out, measuring the half value layers (HVL) recommended by the IEC 61267 standard, after that the RQT 8, RQT 9 and RQT 10 radiation quality references were established. For additional filters, aluminum and copper of high purity (around 99.9%) were used. RQT's in thickness of copper filters equivalent to the set 'RQR (Al) + Additional Filtration (Cu)' was directly found by an alternative methodology used to determine additional filtrations, which is a good option when RQR's have not the possibility of be setting up. With the establishment of this new methodology for the ionization pencil chambers calibration, the LCI is ready to calibrate these instruments according to the most recent international standards. Therefore, an improvement in calibration traceability, as well as in metrological services offered by IPEN to all Brazil is achieved. (author)

  17. Accreditation of testing laboratories in dosimetry: The use of a flexible scope at the competent Incorporation Measuring Body Juelich

    International Nuclear Information System (INIS)

    The accreditation of the Competent Incorporation Measuring Body at Juelich includes incorporation monitoring by means of direct measurements of the body activity as well as by means of indirect determination of the body activity by radiochemical analysis of excreta samples. In both testing areas, it proved to be very useful to have a flexible scope. In particular, the associated freedom in choosing testing procedures supports the continual improvement process of the laboratory. The modification of existing methods as well as the development and introduction of new procedures makes an immediate reaction to changed requirements feasible. At Juelich the use made out of the flexible scope included, e.g. the introduction of mathematical calibration in whole-body counting and the automation of sample preparation in radiochemical analysis. Advantages of the new procedures and modified methods include on the one hand the reduction of processing times, downtimes and hazard potentials on the other hand enhanced detection limits and improved cost-efficiency. In the result, it can be recommended to other qualified testing laboratories to go for a flexible scope. (authors)

  18. Calibration of the Gamma Knife Perfexion using TG-21 and the solid water Leksell dosimetry phantom

    International Nuclear Information System (INIS)

    Purpose: To calibrate a Gamma Knife (GK) Perfexion using TG-21 with updated chamber-dependent values for modern microionization chambers in the new solid water Leksell dosimetry phantom. This work illustrates a calibration method using commercially available equipment, instruments, and an established dosimetry protocol that may be adopted at any GK center, thus reducing the interinstitutional variation in GK calibration. The calibration was verified by three third-party dosimetry checks. In addition, measurements of the relative output factors are presented and compared to available data and the new manufacturer-provided relative output factors yet to be released. Methods: An absolute dose calibration based on the TG-21 formalism, utilizing recently reported phantom material and chamber-dependent factors, was performed using a microionization chamber in a spherical solid water phantom. The result was compared to other calibration protocols based on TG-51. Independent verification of the machine output was conducted through M.D. Anderson Dosimetry Services (MDADS), using thermoluminescent dosimeters (TLDs) in an anthropomorphic head phantom; the Radiological Physics Center (RPC), using TLDs in the standard Elekta ABS plastic calibration phantom (gray phantom), included with the GK; and through a collaborative international calibration survey by the University of Pittsburgh Medical Center (UPMC) using alanine dosimeters, also in the gray phantom. The alanine dosimeters were read by the National Institute of Standards and Technology. Finally, Gafchromic EBT film was used to measure relative output factors and these factors were compared to values reported in the literature as well as new values announced for release by Elekta. The films were exposed in the solid water phantom using an included film insert accessory. Results: Compared to the TG-21 protocol in the solid water phantom, the modified and unmodified TG-51 calibrations resulted in dose rates which were 1

  19. Calibration of activation detectors in a monoenergetic neutron beam. Contribution to criticality dosimetry

    International Nuclear Information System (INIS)

    Activation detectors have been calibrated for critical dosimetry applications. Measurements are made using a monoenergetic neutron flux. 14 MeV neutrons obtained par (D-T) reaction are produced by 150 kV accelerator. Neutron flux determined by different methods leads us to obtain an accuracy better than 6%. The present dosimetric system (Activation Neutron Spectrometer - SNAC) gives few informations in the (10 keV - 2 MeV) energetic range. The system has been improved and modified so that SNAC detectors must be read out by gamma spectrometer

  20. A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices.

    Science.gov (United States)

    Fidanzio, Andrea; Greco, Francesca; Gargiulo, Laura; Cilla, Savino; Sabatino, Domenico; Cappiello, Massimo; Di Felice, Cinzia; Di Castro, Elisabetta; Azario, Luigi; Russo, Mariateresa; Pompei, Luciano; D'Onofrio, Guido; Piermattei, Angelo

    2011-03-01

    A practical and accurate generalized in vivo dosimetry procedure has been implemented for Siemens linacs supplying 6, 10, and 15 MV photon beams, equipped with aSi electronic portal imaging devices (EPIDs). The in vivo dosimetry method makes use of correlation ratios between EPID transit signal, s (t) (0) (TPR,w,L), and phantom mid-plane dose, D (0)(TPR,w,L), as functions of phantom thickness, w, square field dimensions, L, and tissue-phantom ratio TPR(20,10). The s (t) (0) (TPR,w,L) and D (0)(TPR,w,L) values were defined to be independent of the EPID sensitivity and monitor unit calibration, while their dependence on TPR(20,10) was investigated to determine a set of generalized correlation ratios to be used for beams with TPR(20,10) falling in the examined range. This way, other radiotherapy centers can use the method with no need to locally perform the whole set of measurements in solid water phantoms, required to implement it. Tolerance levels for 3D conformal treatments, ranging between ±5 and ±6% according to tumor type and location, were estimated for comparison purposes between reconstructed isocenter dose, D (iso), and treatment planning system (TPS) computed dose D (iso,TPS). Finally a dedicated software, interfaceable with record and verify (R&V) systems used in the centers, was developed to obtain in vivo dosimetry results in less than 2 min after beam delivery.

  1. ESR dosimetry in calibration intercomparisons with high-energy photons and electrons

    Energy Technology Data Exchange (ETDEWEB)

    Olsen, K.J. (University Hospital of Copenhagen, Herlev (Denmark)); Hansen, J.W. (Risoe National Lab., Roskilde (Denmark)); Waligorski, M.P.R. (Institute of Nuclear Physics, Cracow (Poland))

    1989-01-01

    When alanine is exposed to radiation, stable free radicals are produced which may be measured by electron spin resonance (ESR) spectroscopy. Our dosimeters consist of L-{alpha}-alanine mixed with 5% polyvinyl pyrrolidone, compacted in the shape of 2-mm thick cylinders of 4.5 mm diameter. The alanine dosimeters were exposed to 0.25 to 16-MV {sub p} X-ray beams and 6 to 20-MeV electron beams and measured at Riso National Laboratory. Doses were determined by comparison with alanine dosimeters exposed to {sup 60}Co {gamma}-rays calibrated with Fricke dosimetry. At absorbed doses above 10 Gy, the standard deviation for the dose measurements was 1%. Dosimetry comparisons better than 2% at the 95% confidence level are possible. The dosimeters are easy to handle, robust and cheap, and may be read repeatedly. Fading after 100 Gy of {sup 60}Co {gamma}-rays and Linac-produced X-ray and electron beams is less than 2 and 6% in 4 years, respectively. Alanine dosimeters are useful for dosimetry comparisons both for photons and electrons, and the negligible fading make them ideal for documentation of patient doses in radiation therapy. (author).

  2. A method for automating calibration and records management for instrumentation and dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    O`Brien, J.M. Jr.; Rushton, R.O.; Burns, R.E. Jr. [Atlan-Tech, Inc., Roswell, GA (United States)

    1993-12-31

    Current industry requirements are becoming more stringent on quality assurance records and documentation for calibration of instruments and dosimetry. A novel method is presented here that will allow a progressive automation scheme to be used in pursuit of that goal. This concept is based on computer-controlled irradiators that can act as stand-alone devices or be interfaced to other components via a computer local area network. In this way, complete systems can be built with modules to create a records management system to meet the needs of small laboratories or large multi-building calibration groups. Different database engines or formats can be used simply by replacing a module. Modules for temperature and pressure monitoring or shipping and receiving can be added, as well as equipment modules for direct IEEE-488 interface to electrometers and other instrumentation.

  3. Neutron Calibration Facilities of the Irsn Research Laboratory in External Dosimetry

    Science.gov (United States)

    van Ryckeghem, L.; Lacoste, V.; Pelcot, G.; Pochat, J.-L.

    2003-06-01

    The Laboratory of Studies and Research in External Dosimetry (LRDE) associated to the National Office for Metrology (BNM) has to maintain the traceability of the French references for the calibration of neutron dosimeters. The LRDE owns a facility which provides some conventional neutron spectra from sources of 241Am-Be, 252Cf, and (252Cf + D2O)/Cd recommended by ISO standards. These ISO spectra appear not appropriated to simulate some kind of workplace spectra. In order to have similar radiation conditions between the calibration and the use of the device, LRDE has built facilities ("SIGMA" and "CANEL") providing some neutron spectra from thermal to fast energies reproducing those encountered in workplaces.

  4. Calibration curves for biological dosimetry; Curvas de calibracion para dosimetria biologica

    Energy Technology Data Exchange (ETDEWEB)

    Guerrero C, C.; Brena V, M. [ININ, A.P. 18-1027, 11801 Mexico D.F. (Mexico)]. E-mail cgc@nuclear.inin.mx

    2004-07-01

    The generated information by the investigations in different laboratories of the world, included the ININ, in which settles down that certain class of chromosomal leisure it increases in function of the dose and radiation type, has given by result the obtaining of calibrated curves that are applied in the well-known technique as biological dosimetry. In this work is presented a summary of the work made in the laboratory that includes the calibrated curves for gamma radiation of {sup 60} Cobalt and X rays of 250 k Vp, examples of presumed exposure to ionizing radiation, resolved by means of aberration analysis and the corresponding dose estimate through the equations of the respective curves and finally a comparison among the dose calculations in those people affected by the accident of Ciudad Juarez, carried out by the group of Oak Ridge, USA and those obtained in this laboratory. (Author)

  5. Calibration of survey meters at the Algerian Secondary Standard Dosimetry Laboratory

    International Nuclear Information System (INIS)

    Full text: The Algerian Secondary Standard Dosimetry Laboratory, which became a Regional AFRA Designated Centre for French spoken countries, in the field of calibration since 2005, has developed and implemented methodologies for the calibration of radiation protection instruments such as survey meters. These instruments were initially calibrated in terms of air kerma free in air (NK) in 60Co, 137Cs gamma radiations. Although the recommended calibration quantity is H*(10) (Sv/h), most of instruments measure different physical quantities and problems about the expression of calibration factor in the calibration certificates are always raised. In order to switch to calibrations in terms of H*(10), a feasibility study was performed using instruments submitted for calibration at the SSDL. This paper presents the variation of the calibration coefficients, Fc, obtained since 2005 for three types of survey meters. The uncertainty components involved in the determination of Fc are analysed in details. Material and method: This study included the most used survey meters in Algeria and received at the SSDL in 2005 and 2006 which consist of 127 survey-meters of type Automess (different models), 80 Ludlum model 3 and 23 Graetz X 5 DE. Calibrations were performed using the beam output decay method. This output was determined with the Reference Standard chamber of type NE 2575 in 2005 and LS01 in 2006, both calibrated at the IAEA calibration laboratory. The average calibration coefficients, Fc, which converts the instruments reading to ambient equivalent absorbed dose, are calculated and the relative standard deviations assessed. The variation of calibration coefficients, for each type of instruments, is compared to the overall stated uncertainty for this coefficient. Results and discussion: The mean calibration coefficients in terms of H*(10) were lying from 0.993 ± 2.55% and 1.039 ± 3.16% for the Automess survey meters and from . The results are 1.215 ± 5.93%, 1.149 ± 14

  6. A suitability study of the fission product phantom and the bottle manikin absorption phantom for calibration of in vivo bioassay equipment for the DOELAP accreditation testing program

    Energy Technology Data Exchange (ETDEWEB)

    Olsen, P.C.; Lynch, T.P.

    1991-08-01

    Pacific Northwest laboratory (PNL) conducted an intercomparison study of the Fission Product phantom and the bottle manikin absorption (BOMAB) phantom for the US Department of Energy (DOE) to determine the consistency of calibration response of the two phantoms and their suitability for certification and use under a planned bioassay laboratory accreditation program. The study was initiated to determine calibration factors for both types of phantoms and to evaluate the suitability of their use in DOE Laboratory Accreditation Program (DOELAP) round-robin testing. The BOMAB was found to be more appropriate for the DOELAP testing program. 9 refs., 9 figs., 9 tabs.

  7. Optimization of MOSFET-type sensor calibration for the implementation of in vivo dosimetry in stereotactic radiosurgery; Optimisation de la calibration de capteurs de type MOSFET pour la mise en oeuvre de la dosimetrie in vivo en radiochirurgie stereotaxique

    Energy Technology Data Exchange (ETDEWEB)

    Sors, A. [Laboratoire d' etude et de recherche en imagerie spatiale et medicale, universite Paul-Sabatier, 31 - Toulouse (France); Cassol, E.; Duthil, P. [Unite de radiophysique et de radioprotection, CHU Toulouse, 31 (France); Cassol, E.; Lotterie, J.A.; Berry, I.; Franceries, X. [Inserm UMRS 825, 31 - Toulouse (France); Hallil, A. [Best Medical Canada, Ottawa (Canada); Sors, A.; Latorzeff, I.; Lotterie, J.A.; Redon, A.; Berry, I. [Centre de radiochirurgie stereotaxique, CHU Rangueil, 31 - Toulouse (France); Latorzeff, I.; Redon, A.; Berry, I. [Groupe Oncorad Garonne, 31 - Toulouse (France); Sors, A.; Cassol, E.; Hallil, A.; Latorzeff, I.; Duthil, P.; Lotterie, J.A.; Redon, A.; Berry, I.; Franceries, X. [Universite Paul-Sabatier, Toulouse-3, 31 - Toulouse (France)

    2010-10-15

    Within the frame of a project of assessment of in vivo dosimetry methods in stereotactic radiosurgery delivering an irradiation by a conformational dynamic arc therapy technique, the authors more precisely report the assessment and optimization of the calibration of MOSFET and micro MOSFET sensors. Measurements are performed on a Navel's system equipped with a multi-blade collimator. Short communication

  8. A global calibration model for a-Si EPIDs used for transit dosimetry.

    Science.gov (United States)

    Nijsten, S M J J G; van Elmpt, W J C; Jacobs, M; Mijnheer, B J; Dekker, A L A J; Lambin, P; Minken, A W H

    2007-10-01

    Electronic portal imaging devices (EPIDs) are not only applied for patient setup verification and detection of organ motion but are also increasingly used for dosimetric verification. The aim of our work is to obtain accurate dose distributions from a commercially available amorphous silicon (a-Si) EPID for transit dosimetry applications. For that purpose, a global calibration model was developed, which includes a correction procedure for ghosting effects, field size dependence and energy dependence of the a-Si EPID response. In addition, the long-term stability and additional buildup material for this type of EPID were determined. Differences in EPID response due to photon energy spectrum changes have been measured for different absorber thicknesses and field sizes, yielding off-axis spectrum correction factors based on transmission measurements. Dose measurements performed with an ionization chamber in a water tank were used as reference data, and the accuracy of the dosimetric calibration model was determined for a large range of treatment conditions. Gamma values using 3% as dose-difference criterion and 3 mm as distance-to-agreement criterion were used for evaluation. The field size dependence of the response could be corrected by a single kernel, fulfilling the gamma evaluation criteria in case of virtual wedges and intensity modulated radiation therapy fields. Differences in energy spectrum response amounted up to 30%-40%, but could be reduced to less than 3% using our correction model. For different treatment fields and (in)homogeneous phantoms, transit dose distributions satisfied in almost all situations the gamma criteria. We have shown that a-Si EPIDs can be accurately calibrated for transit dosimetry purposes. PMID:17985633

  9. DOSIMETRY

    CERN Multimedia

    2001-01-01

    From the month of May on, the neutron dosimeter will be worn in an extra package distinct from the usual film-badge. We will give you more ample information in Weekly Bulletin No. 18/2001 of April 30, 2001. In the week following Easter (17 - 20. 4. 2001) the Individual Dosimetry Service will be opened in the mornings from 8:30 to 11:30 h only. The Service will be closed on April 30.

  10. Methods of measurement and calibration in personnel dosimetry for external irradiation: presentation of the concept and the results of a test programme in Switzerland

    International Nuclear Information System (INIS)

    In 1986 the Swiss working group on personnel dosimetry issued a recommendation on methods of calibration and measurement in personnel dosimetry for external radiation, based on ICRU report 39. The aim of the report was to give detailed and pragmatic guidance to the dosimetry services and the calibration labs. In the recommendation a definition is given for primary and secondary limited quantities, for calibration quantities and operational quantities as well as for their interrelationships. The second part of the report is a manual for the calibration and evaluation procedures for whole-body and extremity dosemeters for photon and electron radiation. Based on that recommendation an extended test programme has been performed for all dosimetry services in Switzerland. The results show that the legal introduction of the recommended concept has no severe consequences for the dosimetry services although a number of minor modifications are needed. (author)

  11. Dose levels of the occupational radiation exposures in Poland based on results from the accredited dosimetry service at the IFJ PAN, Krakow.

    Science.gov (United States)

    Budzanowski, Maciej; Kopeć, Renata; Obryk, Barbara; Olko, Paweł

    2011-03-01

    Individual dosimetry service based on thermoluminescence (TLD) detectors has started its activity at the Institute of Nuclear Physics (IFJ) in Krakow in 1965. In 2002, the new Laboratory of Individual and Environment Dosimetry (Polish acronym LADIS) was established and underwent the accreditation according to the EN-PN-ISO/IEC 17025 standard. Nowadays, the service is based on the worldwide known standard thermoluminescent detectors MTS-N (LiF:Mg,Ti) and MCP-N (LiF:Mg,Cu,P), developed at IFJ, processed in automatic thermoluminescent DOSACUS or RE2000 (Rados Oy, Finland) readers. Laboratory provides individual monitoring in terms of personal dose equivalent H(p)(10) and H(p)(0.07) in photon and neutron fields, over the range from 0.1 mSv to 1 Sv, and environmental dosimetry in terms of air kerma K(a) over the range from 30 μGy to 1 Gy and also ambient dose equivalent H*(10) over the range from 30 μSv to 1 Sv. Dosimetric service is currently performed for ca. 3200 institutions from Poland and abroad, monitored on quarterly and monthly basis. The goal of this paper is to identify the main activities leading to the highest radiation exposures in Poland. The paper presents the results of statistical evaluation of ∼ 100,000 quarterly H(p)(10) and K(a) measurements performed between 2002 and 2009. Sixty-five per cent up to 90 % of all individual doses in Poland are on the level of natural radiation background. The dose levels between 0.1 and 5 mSv per quarter are the most frequent in nuclear medicine, veterinary and industrial radiography sectors. PMID:21183549

  12. Neutron personnel dosimetry

    International Nuclear Information System (INIS)

    The current state-of-the-art in neutron personnel dosimetry is reviewed. Topics covered include dosimetry needs and alternatives, current dosimetry approaches, personnel monitoring devices, calibration strategies, and future developments

  13. Radiation dosimetry

    CERN Document Server

    Hine, Gerald J; Hine, Gerald J

    1956-01-01

    Radiation Dosimetry focuses on the advancements, processes, technologies, techniques, and principles involved in radiation dosimetry, including counters and calibration and standardization techniques. The selection first offers information on radiation units and the theory of ionization dosimetry and interaction of radiation with matter. Topics include quantities derivable from roentgens, determination of dose in roentgens, ionization dosimetry of high-energy photons and corpuscular radiations, and heavy charged particles. The text then examines the biological and medical effects of radiation,

  14. THE CHALLENGE OF CIEMAT INTERNAL DOSIMETRY SERVICE FOR ACCREDITATION ACCORDING TO ISO/IEC 17025 STANDARD, FOR IN VIVO AND IN VITRO MONITORING AND DOSE ASSESSMENT OF INTERNAL EXPOSURES.

    Science.gov (United States)

    Lopez, M A; Martin, R; Hernandez, C; Navarro, J F; Navarro, T; Perez, B; Sierra, I

    2016-09-01

    The accreditation of an Internal Dosimetry Service (IDS) according to ISO/IEC 17025 Standard is a challenge. The aim of this process is to guarantee the technical competence for the monitoring of radionuclides incorporated in the body and for the evaluation of the associated committed effective dose E(50). This publication describes the main accreditation issues addressed by CIEMAT IDS regarding all the procedures involving good practice in internal dosimetry, focussing in the difficulties to ensure the traceability in the whole process, the appropriate calculation of detection limit of measurement techniques, the validation of methods (monitoring and dose assessments), the description of all the uncertainty sources and the interpretation of monitoring data to evaluate the intake and the committed effective dose. PMID:26433182

  15. A new standard cylindrical graphite-walled ionization chamber for dosimetry in 60Co beams at calibration laboratories

    Science.gov (United States)

    Neves, Lucio P.; Perini, Ana P.; Caldas, Linda V. E.

    2014-11-01

    60Co sources are used mostly at dosimetry laboratories for calibration of ionization chambers utilized for radiotherapy dosimetry, mainly in those laboratories where there is no linear accelerator available. In this work, a new cylindrical ionization chamber was developed and characterized to be used as a reference dosimeter at the Calibration Laboratory of the IPEN. The characterization tests were performed according to the IEC 60731 standard, and all tests presented results within its recommended limits. Furthermore, the correction factors for the wall, stem, central collecting electrode, nonaxial uniformity and the mass-energy absorption coefficient were determined using the EGSnrc Monte Carlo code. The air kerma rate determined with this new dosimeter was compared to the one obtained with the IPEN standard, presenting a difference of 1.5%. Therefore, the new ionization chamber prototype developed and characterized in this work presents potential use as a primary standard dosimeter at radiation metrology laboratories.

  16. Principles for the design and calibration of radiation protection dosemeters for operational and protection quantities for eye lens dosimetry

    International Nuclear Information System (INIS)

    The work package two of the ORAMED project-Collaborative Project (2008-2011) supported by the European Commission within its seventh Framework Programme-is devoted to the study of the eye lens dosimetry. A first approach is to implement the use of Hp(3) by providing new sets of conversion coefficients and well suited calibration and type test procedures. This approach is presented in other papers in the proceedings of this conference. Taking into account that the eye lens is an organ close to the surface of the body, another approach would be to directly estimate the absorbed dose to the eye lens, Dlens,est through a special calibration procedure although this quantity is not directly measurable. This paper is a methodological paper that tries to identify the critical aspects of a dosimetry in terms of Dlens. (authors)

  17. Principles for the design and calibration of radiation protection dosemeters for operational and protection quantities for eye lens dosimetry.

    Science.gov (United States)

    Bordy, J M; Gualdrini, G; Daures, J; Mariotti, F

    2011-03-01

    The work package two of the ORAMED project--Collaborative Project (2008-2011) supported by the European Commission within its seventh Framework Programme--is devoted to the study of the eye lens dosimetry. A first approach is to implement the use of H(p)(3) by providing new sets of conversion coefficients and well suited calibration and type test procedures. This approach is presented in other papers in the proceedings of this conference. Taking into account that the eye lens is an organ close to the surface of the body, another approach would be to directly estimate the absorbed dose to the eye lens, D(lens,est) through a special calibration procedure although this quantity is not directly measurable. This paper is a methodological paper that tries to identify the critical aspects of a dosimetry in terms of D(lens).

  18. Optimization of MOSFET calibration for in vivo dosimetry in radiosurgery: reduction of measurement uncertainties in pre-clinical conditions; Optimisation de la calibration de MOSFET pour la dosimetrie in vivo en radiochirurgie: reduction des incertitudes de mesure en conditions precliniques

    Energy Technology Data Exchange (ETDEWEB)

    Sors, A.; Berry, I.; Franceries, X. [UMR 825 ' imagerie cerebrale et handicaps neurologiques' , Inserm, Toulouse (France); Cassol, E.; Duthil, P. [Unite de radiophysique et de radioprotection, CHU de Toulouse, Toulouse (France); Hallil, A. [Best medical Canada, Ottawa (Canada); Latorzeff, I.; Lotterie, J.A. [Centre de radiochirurgie stereotaxique, CHU Rangueil, Toulouse (France); Redon, A. [Groupe Oncorad Garonne, Montauban (France)

    2011-10-15

    The objective of this study is to assess the conventional formulas of equivalent square for fields with irregular geometry, by transposing the optimized calibration method which has been previously developed, to micro-MOSFET. The study has been performed on a 6 MV Novalis apparatus equipped with micro-multi-blades collimators (BrainLab). The average dose bias reaches 2.66 per cent for all field sizes. Therefore, it appears that the joint use of the square inverse of distances and of conventional formulas of equivalent square results in an acceptable in vivo dosimetry precision. Short communication

  19. Biological dosimetry of ionizing radiation: Evaluation of the dose with cytogenetic methodologies by the construction of calibration curves

    Science.gov (United States)

    Zafiropoulos, Demetre; Facco, E.; Sarchiapone, Lucia

    2016-09-01

    In case of a radiation accident, it is well known that in the absence of physical dosimetry biological dosimetry based on cytogenetic methods is a unique tool to estimate individual absorbed dose. Moreover, even when physical dosimetry indicates an overexposure, scoring chromosome aberrations (dicentrics and rings) in human peripheral blood lymphocytes (PBLs) at metaphase is presently the most widely used method to confirm dose assessment. The analysis of dicentrics and rings in PBLs after Giemsa staining of metaphase cells is considered the most valid assay for radiation injury. This work shows that applying the fluorescence in situ hybridization (FISH) technique, using telomeric/centromeric peptide nucleic acid (PNA) probes in metaphase chromosomes for radiation dosimetry, could become a fast scoring, reliable and precise method for biological dosimetry after accidental radiation exposures. In both in vitro methods described above, lymphocyte stimulation is needed, and this limits the application in radiation emergency medicine where speed is considered to be a high priority. Using premature chromosome condensation (PCC), irradiated human PBLs (non-stimulated) were fused with mitotic CHO cells, and the yield of excess PCC fragments in Giemsa stained cells was scored. To score dicentrics and rings under PCC conditions, the necessary centromere and telomere detection of the chromosomes was obtained using FISH and specific PNA probes. Of course, a prerequisite for dose assessment in all cases is a dose-effect calibration curve. This work illustrates the various methods used; dose response calibration curves, with 95% confidence limits used to estimate dose uncertainties, have been constructed for conventional metaphase analysis and FISH. We also compare the dose-response curve constructed after scoring of dicentrics and rings using PCC combined with FISH and PNA probes. Also reported are dose response curves showing scored dicentrics and rings per cell, combining

  20. CT head-scan dosimetry in an anthropomorphic phantom and associated measurement of ACR accreditation-phantom imaging metrics under clinically representative scan conditions

    Energy Technology Data Exchange (ETDEWEB)

    Brunner, Claudia C.; Stern, Stanley H.; Chakrabarti, Kish [U.S. Food and Drug Administration, 10903 New Hampshire Avenue, Silver Spring, Maryland 20993 (United States); Minniti, Ronaldo [National Institute of Standards and Technology, 100 Bureau Drive, Gaithersburg, Maryland 20899 (United States); Parry, Marie I. [Walter Reed National Military Medical Center, 8901 Rockville Pike, Bethesda, Maryland 20889 (United States); Skopec, Marlene [National Institutes of Health, 9000 Rockville Pike, Bethesda, Maryland 20892 (United States)

    2013-08-15

    Purpose: To measure radiation absorbed dose and its distribution in an anthropomorphic head phantom under clinically representative scan conditions in three widely used computed tomography (CT) scanners, and to relate those dose values to metrics such as high-contrast resolution, noise, and contrast-to-noise ratio (CNR) in the American College of Radiology CT accreditation phantom.Methods: By inserting optically stimulated luminescence dosimeters (OSLDs) in the head of an anthropomorphic phantom specially developed for CT dosimetry (University of Florida, Gainesville), we measured dose with three commonly used scanners (GE Discovery CT750 HD, Siemens Definition, Philips Brilliance 64) at two different clinical sites (Walter Reed National Military Medical Center, National Institutes of Health). The scanners were set to operate with the same data-acquisition and image-reconstruction protocols as used clinically for typical head scans, respective of the practices of each facility for each scanner. We also analyzed images of the ACR CT accreditation phantom with the corresponding protocols. While the Siemens Definition and the Philips Brilliance protocols utilized only conventional, filtered back-projection (FBP) image-reconstruction methods, the GE Discovery also employed its particular version of an adaptive statistical iterative reconstruction (ASIR) algorithm that can be blended in desired proportions with the FBP algorithm. We did an objective image-metrics analysis evaluating the modulation transfer function (MTF), noise power spectrum (NPS), and CNR for images reconstructed with FBP. For images reconstructed with ASIR, we only analyzed the CNR, since MTF and NPS results are expected to depend on the object for iterative reconstruction algorithms.Results: The OSLD measurements showed that the Siemens Definition and the Philips Brilliance scanners (located at two different clinical facilities) yield average absorbed doses in tissue of 42.6 and 43.1 m

  1. Dosimetry in high-energy photon fields for the calibration of measuring instruments for radiation protection purposes

    International Nuclear Information System (INIS)

    This report describes the dosimetry in various reference photon fields with energies between 4.4 MeV and about 8 MeV. Two dosimetric quantities were chosen. The air kerma was determined from measurements without a phantom and the absorbed dose to water from measurements with a phantom. This mean that the range of realization of the quantity air kerma has been extended from the energy of Co-60 photons to about 8 MeV. The results can serve as basis for the calibration of radiation protection dosemeters in nuclear power plants (0-16(n,p)N-16 reaction) with high energy photons. (orig./HP)

  2. Field calibration of PADC track etch detectors for local neutron dosimetry in man using different radiation qualities

    Energy Technology Data Exchange (ETDEWEB)

    Haelg, Roger A., E-mail: rhaelg@phys.ethz.ch [Institute for Radiotherapy, Radiotherapie Hirslanden AG, Hirslanden Medical Center, Rain 34, CH-5000 Aarau (Switzerland); Besserer, Juergen [Institute for Radiotherapy, Radiotherapie Hirslanden AG, Hirslanden Medical Center, Rain 34, CH-5000 Aarau (Switzerland); Boschung, Markus; Mayer, Sabine [Division for Radiation Safety and Security, Paul Scherrer Institut, CH-5232 Villigen (Switzerland); Clasie, Benjamin [Department of Radiation Oncology, Massachusetts General Hospital, 30 Fruit Street, Boston, MA 02114 (United States); Kry, Stephen F. [Department of Radiation Physics, The University of Texas M.D. Anderson Cancer Center, 1515 Holcombe Blvd., Houston, TX 77030 (United States); Schneider, Uwe [Institute for Radiotherapy, Radiotherapie Hirslanden AG, Hirslanden Medical Center, Rain 34, CH-5000 Aarau (Switzerland); Vetsuisse Faculty, University of Zurich, Winterthurerstrasse 204, CH-8057 Zurich (Switzerland)

    2012-12-01

    In order to quantify the dose from neutrons to a patient for contemporary radiation treatment techniques, measurements inside phantoms, representing the patient, are necessary. Published reports on neutron dose measurements cover measurements performed free in air or on the surface of phantoms and the doses are expressed in terms of personal dose equivalent or ambient dose equivalent. This study focuses on measurements of local neutron doses inside a radiotherapy phantom and presents a field calibration procedure for PADC track etch detectors. An initial absolute calibration factor in terms of H{sub p}(10) for personal dosimetry is converted into neutron dose equivalent and additional calibration factors are derived to account for the spectral changes in the neutron fluence for different radiation therapy beam qualities and depths in the phantom. The neutron spectra used for the calculation of the calibration factors are determined in different depths by Monte Carlo simulations for the investigated radiation qualities. These spectra are used together with the energy dependent response function of the PADC detectors to account for the spectral changes in the neutron fluence. The resulting total calibration factors are 0.76 for a photon beam (in- and out-of-field), 1.00 (in-field) and 0.84 (out-of-field) for an active proton beam and 1.05 (in-field) and 0.91 (out-of-field) for a passive proton beam, respectively. The uncertainty for neutron dose measurements using this field calibration method is less than 40%. The extended calibration procedure presented in this work showed that it is possible to use PADC track etch detectors for measurements of local neutron dose equivalent inside anthropomorphic phantoms by accounting for spectral changes in the neutron fluence.

  3. Radioluminescence (RL) probe dosimetry using Al2O3:C for precise calibration of beta sources applied in luminescence dating

    International Nuclear Information System (INIS)

    The radioluminescence (RL) phenomenon is widely known as prompt luminescence or fluorescence of synthetic materials, like Al2O3 and LiF as well as of natural dosimeter materials, like feldspar and quartz, during interaction with ionising radiation, especially electron rays. Almost all of these materials show increasing RL flux with increasing absorbed beta dose. Only potassium rich feldspars (microcline and orthoclase) show a RL emission at 1.45 eV of decreasing intensity with rising delivered dose. For a number of reasons, this RL emission is suitable for dating purposes. Accuracy and reproducibility of the RL dating method are much more advantageous compared to TL and OSL methods. Carbon doped Al2O3 (TLD 500) shows a much more higher RL intensity than, e.g., feldspar and the stability of the stored electrons as well as the linearity of RL yield versus primary electron energy, qualifies this dosimeter material to be a useful tool in source calibration, using radioluminescence. However, the accuracy of a luminescence dating method depends largely on the calibration of the used irradiation sources. To prevent calibration uncertainties due to the use of natural dosimeters with a high scattering of luminescence yield, a new method is suggested. This method bases on probe dosimetry using a very small amount of Al2O3:C as probe and, consequently, the material that should be calibrated (e.g. feldspar). The amount of Al2O3:C may disturb the electron fluence within the material to be calibrated only insignificantly. The sample mix is irradiated by an exactly known Gamma or Beta radiation source and the absorbed dose is treated as absorbed dose in the appropriated material. In contrast to conventional calibration methods, the suggested performance uses the dose dependent RL signal of Al2O3:C at 3.00 eV. This shows a good reproducibility, even with a tiny amount of the applied synthetic probe material. (author)

  4. Calibration of semiconductors diodes for in vivo dosimetry in total body irradiation treatments; Calibracao de diodos semicondutores para dosimetria in vivo em tratamentos de irradiacao de corpo inteiro

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Fernanda F.; Costa, Alessandro M.; Ghilardi Netto, Thomaz, E-mail: ferretti.oliveira@gmail.com [Universidade de Sao Paulo (FFCLRP/USP), Ribeirao Preto, SP (Brazil). Faculdade de Ciencias e Letras. Departamento de Fisica; Amaral, Leonardo L. [Universidade de Sao Paulo (HCFMRP/USP), Ribeirao Preto, SP (Brazil). Hospital das Clinicas. Servico de Radioterapia

    2012-08-15

    This paper presents the results of in vivo dosimetry with p-type semiconductors diodes, EDP-15 (Scanditronix Wellhoefer) of two patients who underwent total body irradiation treatments, at Hospital das Clinicas da Faculdade de Medicina de Ribeirao Preto University of Sao Paulo (HCFMRP-USP). The diodes were well calibrated and the calibration factors were determined with the aid of a reference ionization chamber (FC065, IBA dosimetry, sensitive volume of 0.65 cm{sup 3}).The calibration was performed in a Total Body Irradiation (TBI) setup, using solid water phantoms. Different lateral thicknesses from one patient were simulated and then the calibration factors were determined by means of maximum depth dose readings (half of the lateral thickness). The response difference between diode readings and the prescribed dose for both treatments was below 4%. This difference is in agreement as recommended by International Commission on Radiation Units (ICRU), which is {+-}5%. (author)

  5. H p(0.07) photon dosemeters for eye lens dosimetry: Calibration on a rod vs. a slab phantom

    International Nuclear Information System (INIS)

    In recent years, several papers dealing with eye lens dosimetry have been published as epidemiological studies are implying that the induction of cataracts occurs even at eye lens doses of less than 500 mGy. For that reason, the necessity to monitor the eye lens may become more important than it was before. However, only few dosemeters for the appropriate quantity Hp(3) are available. Partial-body dosemeters are usually designed to measure the quantity Hp(0.07) calibrated on a rod phantom representing a finger while a slab phantom much better represents the head. Therefore, in this work it was investigated whether dosemeters designed for the quantity Hp(0.07) calibrated on a rod phantom can also be worn on the head (close to the eyes) and still deliver correct results (Hp(0.07) on a head). For that purpose, different types of partial-body dosemeters from routine use were irradiated at different photon energies on both a rod and a slab phantom. It turned out that their response values are within ±5% independent of the phantom if the quantity value for the respective phantom is used. Thus, partial-body dosemeters designed for the quantity Hp(0.07) calibrated on a rod phantom may be worn on the head and used to monitor the eye lens dose due to photon radiation via the measurement of Hp(0.07) on the head. (authors)

  6. Improvement of the WBC calibration of the Internal Dosimetry Laboratory of the CDTN/CNEN using MCNPX code

    Energy Technology Data Exchange (ETDEWEB)

    Guerra P, F.; Heeren de O, A. [Universidade Federal de Minas Gerais, Departamento de Engenharia Nuclear, Programa de Pos Graduacao em Ciencias e Tecnicas Nucleares, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil); Melo, B. M.; Lacerda, M. A. S.; Da Silva, T. A.; Ferreira F, T. C., E-mail: tcff01@gmail.com [Centro de Desenvolvimento da Tecnologia Nuclear, Programa de Pos Graduacao / CNEN, Av. Pte. Antonio Carlos 6627, 31270-901 Belo Horizonte, Minas Gerais (Brazil)

    2015-10-15

    The Plan of Radiological Protection licensed by the National Nuclear Energy Commission - CNEN in Brazil includes the risks of assessment of internal and external exposure by implementing a program of individual monitoring which is responsible of controlling exposures and ensuring the maintenance of radiation safety. The Laboratory of Internal Dosimetry of the Center for Development of Nuclear Technology - LID/CDTN is responsible for routine monitoring of internal contamination of the Individuals Occupationally Exposed (IOEs). These are, the IOEs involved in handling {sup 18}F produced by the Unit for Research and Production of Radiopharmaceuticals sources; as well a monitoring of the entire body of workers from the Research Reactor TRIGA IPR-R1/CDTN or whenever there is any risk of accidental incorporation. The determination of photon emitting radionuclides from the human body requires calibration techniques of the counting geometries, in order to obtain a curve of efficiency. The calibration process normally makes use of physical phantoms containing certified activities of the radionuclides of interest. The objective of this project is the calibration of the WBC facility of the LID/CDTN using the BOMAB physical phantom and Monte Carlo simulations. Three steps were needed to complete the calibration process. First, the BOMAB was filled with a KCl solution and several measurements of the gamma ray energy (1.46 MeV) emitted by {sup 40}K were done. Second, simulations using MCNPX code were performed to calculate the counting efficiency (Ce) for the BOMAB model phantom and compared with the measurements Ce results. Third and last step, the modeled BOMAB phantom was used to calculate the Ce covering the energy range of interest. The results showed a good agreement and are within the expected ratio between the measured and simulated results. (Author)

  7. Dosimetry standards

    International Nuclear Information System (INIS)

    The following leaflets are contained in this folder concerning the National Physical Laboratory's measurement services available in relation to dosimetry standards: Primary standards of X-ray exposure and X-ray irradiation facilities, X-ray dosimetry at therapy levels, Protection-level X-ray calibrations, Therapy-level gamma-ray facility, Fricke dosemeter reference service, Low-dose-rate gamma-ray facility, Penetrameter and kV meter calibration, Measurement services for radiation processing, Dichromate dosemeter reference service, Electron linear accelerator. (U.K.)

  8. A reference current source for the calibration of current measuring systems in dosimetry using ionisation chambers

    International Nuclear Information System (INIS)

    This work describes a newly developed, easily handled current source containing one 90Sr + 90Y Y beta emitter. Four different currents from 7x10-11A to 4x10-14A are supplied. From 5 to 10 control measurements were performed for each of the four currents during a period of eight months. The standard deviation of the values of these control measurements, corrected to a reference air density and for radioactive decay, was between 0,07% and 0,4% with respect to the mean values obtained during the eight months. The relative standard deviation of the single values belonging to a control measurement carried out on one day under constant ambient conditions amounts to approximately one fourth of the values quoted above. The current source was utilized in a number of laboratories in the PTB and proved its reliability for checking current measuring systems in dosimetry with ionization chambers. (orig.)

  9. Ophthalmic applicators: An overview of calibrations following the change to SI units

    International Nuclear Information System (INIS)

    Since the NIST dose to water standard for 90Sr/90Y ophthalmic applicators was introduced, numerous sources have undergone calibration either at NIST or at the University of Wisconsin Accredited Dosimetry Calibration Laboratory (UWADCL). From 1997 to 2008, 222 of these beta-emitting sources were calibrated at the UWADCL, and prior reference source strength values were available for 149 of these sources. A survey of UWADCL ophthalmic applicator calibrations is presented here, demonstrating an average discrepancy of -19% with a standard deviation of ±16% between prior reference values and the NIST-traceable UWADCL absorbed dose to water calibrations. Values ranged from -49% to +42%.

  10. DOE standard: The Department of Energy Laboratory Accreditation Program administration

    International Nuclear Information System (INIS)

    This technical standard describes the US Department of Energy Laboratory Accreditation Program (DOELAP), organizational responsibilities, and the accreditation process. DOELAP evaluates and accredits personnel dosimetry and radiobioassay programs used for worker monitoring and protection at DOE and DOE contractor sites and facilities as required in Title 10, Code of Federal Regulations, Part 835, Occupational Radiation Protection. The purpose of this technical standard is to establish procedures for administering DOELAP and acquiring accreditation

  11. Accreditation of environmental TLD dosimetry system: validation and uncertainty of the method; Acreditacion de sistema de dosimetria ambiental TLD: validacion e incertumbre del metodo

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez Jimenez, R.; Romero Gutierrez, A. M.; Lopez Moyano, J. L.

    2013-07-01

    he work shows the results obtained in the validation of the method used, and the calculation of uncertainty. The authors' goal is to spread the practical experience gained during the accreditation process, paying special attention to the process of validation of the method and the estimation of the uncertainty of the dosimetric systems. (Author)

  12. Dosimetry through the Secondary Laboratory of Dosimetric Calibration of Mexico; Dosimetria a traves del Laboratorio Secundario de Calibracion Dosimetrica de Mexico

    Energy Technology Data Exchange (ETDEWEB)

    Tovar M, V.M.; Alvarez R, J.T.; Medina O, V.P.; Vergara M, F.; Anaya M, R.; Cejudo A, J.; Salinas L, B. [ININ, A.P. 18-1027, 11801 Mexico D.F. (Mexico)

    2004-07-01

    In the beginnings of the sixty years an urgent necessity is presented mainly in the developing countries, of improving in important form the accuracy in the dosimetry of external faces in therapy of radiations (radiotherapy centers), mainly in the calibration of ''clinical dosemeters''. In 1976 the International Atomic Energy Agency, (IAEA), and the World Health Organization, (WHO), they carried out a mutual agreement with regard to the establishment and operation of a net of Secondary Patron Laboratories of Dosimetry, (LSCD). The necessity to establish measure patterns in the field of the dosimetry of the ionizing radiations, is necessary, to have an accuracy but high in the dosimetry of the radiation beams in therapy which is highly dependent of the dose given to the tumor of those patient with cancer. Similar levels of accuracy are required in protection measures to the radiation with an acceptable smaller accuracy, however, when the personal dosemeters are used to determine the doses received by the individuals under work conditions, such mensurations in therapy of radiations and radiological protection will have traceability through a chain of comparisons to primary or national patterns. The traceability is necessary to assure the accuracy and acceptability of the dosimetric measures, as well as, the legal and economic implications. The traceability is also necessary in the dosimetry of high dose like in the sterilization of different products. The main function of the LSCD is to provide a service in metrology of ionizing radiations, maintaining the secondary or national patterns, which have a traceability to the International System of measures, which is based for if same in the comparison of patterns in the Primary Laboratories of Dosimetry (LPD) under the auspice of the International Office of Weights and Measure (BIPM). The secondary and national patterns in the LSCD constitute in Mexico, the national patterns of the magnitudes in the

  13. NVLAP activities at Department of Defense calibration laboratories

    Energy Technology Data Exchange (ETDEWEB)

    Schaeffer, D.M. [Defense Nuclear Agency, Alexandria, VA (United States)

    1993-12-31

    There are 367 active radiological instrument calibration laboratories within the U.S. Department of Defense (DoD). Each of the four services in DoD manages, operates, and certifies the technical proficiency and competency of those laboratories under their cognizance. Each service has designated secondary calibration laboratories to trace all calibration source standards to the National Institute of Standards and Technology. Individual service radiological calibration programs and capabilities, present and future, are described, as well as the measurement quality assurance (MQA) processes for their traceability. National Voluntary Laboratory Accreditation Program (NVLAP) programs for dosimetry systems are briefly summarized. Planned NVLAP accreditation of secondary laboratories is discussed in the context of current technical challenges and future efforts.

  14. Development of dose calibrators Tandem systems and establishment of beta dosimetry in nuclear medicine

    International Nuclear Information System (INIS)

    A quality control program at Nuclear Medicine Services includes the checking of all equipment used for diagnostics and treatment, and the individual monitoring of the workers occupationally exposed to ionizing radiations. In this work the main quality control tests were performed with three dose calibrators using standard radiation sources of 57Co, 133Ba, 137Cs and 60Co. Tandem systems of dose calibrators were established and characterized using four cylindrical absorbers of different materials for an additional quality control test in Nuclear Medicine. The main utility of this new test is the possibility of impurity detection in radiopharmaceuticals, when the ratio of the measurements with different absorbers is different from that obtained at the laboratory in ideal conditions. The dosimetric characteristics of three types of CaS04:Dy + Teflon pellets were studied for an appropriate choice of the material to be used for individual monitoring of workers. The thermoluminescent detectors were irradiated using beta sources of 90Sr+90Y, 204TI, 147Pm, 153Sm and 32P. A wrist badge for beta individual monitoring was developed for workers that handle beta radiopharmaceuticals in Nuclear Medicine Services. (author)

  15. Clinical laboratory accreditation in India.

    Science.gov (United States)

    Handoo, Anil; Sood, Swaroop Krishan

    2012-06-01

    Test results from clinical laboratories must ensure accuracy, as these are crucial in several areas of health care. It is necessary that the laboratory implements quality assurance to achieve this goal. The implementation of quality should be audited by independent bodies,referred to as accreditation bodies. Accreditation is a third-party attestation by an authoritative body, which certifies that the applicant laboratory meets quality requirements of accreditation body and has demonstrated its competence to carry out specific tasks. Although in most of the countries,accreditation is mandatory, in India it is voluntary. The quality requirements are described in standards developed by many accreditation organizations. The internationally acceptable standard for clinical laboratories is ISO15189, which is based on ISO/IEC standard 17025. The accreditation body in India is the National Accreditation Board for Testing and Calibration Laboratories, which has signed Mutual Recognition Agreement with the regional cooperation the Asia Pacific Laboratory Accreditation Cooperation and with the apex cooperation the International Laboratory Accreditation Cooperation. PMID:22727005

  16. Quality assurance in radiotherapy dosimetry in China

    International Nuclear Information System (INIS)

    In 1995, the SSDL in the Laboratory of Industrial Hygiene cooperated with Beijing Cancer Hospital, Chinese Academy of Medical science joined the IAEA Co-ordinated Research Programme (NO.8769/RO). According to the requirements of the project, an External Audit Group (EAG) in China was established in 1996 with the responsibilities of operating TLD-based quality audit for radiotherapy dosimetry. Since then. The national TLD dose quality audit services have been carried out in 7 provinces in China. Besides this, the national programmes for brachytherapy and stereostatic radiosurgery (SRS) treatment dosimetry were initiated in 2001. The activity measurement intercomparison between the SSDL and some hospitals for Ir-192 HDR brachytherapy sources has been performed using a HDR well-type ionization chamber (Model HDR 1000 plus) and CDX-2000A Charge Digitizer, which were calibrated in Accredited Dosimetry Calibration Laboratory, University of Wisconsin, USA. The preliminary results indicated that the agreement between SSDL measured activity and hospital stated activity was within ±5% for more than 80% of total participants

  17. Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure

    International Nuclear Information System (INIS)

    Film dosimetry is an attractive tool for dose distribution verification in intensity modulated radiotherapy (IMRT). A critical aspect of radiochromic film dosimetry is the scanner used for the readout of the film: the output needs to be calibrated in dose response and corrected for pixel value and spatial dependent nonuniformity caused by light scattering; these procedures can take a long time. A method for a fast and accurate calibration and uniformity correction for radiochromic film dosimetry is presented: a single film exposure is used to do both calibration and correction. Gafchromic EBT films were read with two flatbed charge coupled device scanners (Epson V750 and 1680Pro). The accuracy of the method is investigated with specific dose patterns and an IMRT beam. The comparisons with a two-dimensional array of ionization chambers using a 18x18 cm2 open field and an inverse pyramid dose pattern show an increment in the percentage of points which pass the gamma analysis (tolerance parameters of 3% and 3 mm), passing from 55% and 64% for the 1680Pro and V750 scanners, respectively, to 94% for both scanners for the 18x18 open field, and from 76% and 75% to 91% for the inverse pyramid pattern. Application to an IMRT beam also shows better gamma index results, passing from 88% and 86% for the two scanners, respectively, to 94% for both. The number of points and dose range considered for correction and calibration appears to be appropriate for use in IMRT verification. The method showed to be fast and to correct properly the nonuniformity and has been adopted for routine clinical IMRT dose verification

  18. Fourth conference on radiation protection and dosimetry: Proceedings, program, and abstracts

    Energy Technology Data Exchange (ETDEWEB)

    Casson, W.H.; Thein, C.M.; Bogard, J.S. [eds.

    1994-10-01

    This Conference is the fourth in a series of conferences organized by staff members of Oak Ridge National Laboratory in an effort to improve communication in the field of radiation protection and dosimetry. Scientists, regulators, managers, professionals, technologists, and vendors from the United States and countries around the world have taken advantage of this opportunity to meet with their contemporaries and peers in order to exchange information and ideas. The program includes over 100 papers in 9 sessions, plus an additional session for works in progress. Papers are presented in external dosimetry, internal dosimetry, radiation protection programs and assessments, developments in instrumentation and materials, environmental and medical applications, and on topics related to standards, accreditation, and calibration. Individual papers are indexed separately on EDB.

  19. Measurement assurance in dosimetry

    International Nuclear Information System (INIS)

    The uses of radiation in medicine and industry are today wide in scope and diversity and there is a need for reliable dosimetry in most applications. In particular, high accuracy in dosimetry is required in the therapeutic use of radiation. Consequently, calibration procedures for radiotherapy generally meet also the accuracy requirements for applications in other fields, such as diagnostic radiology, radiation protection and industrial radiation processing. The emphasis at this symposium was therefore mainly or radiotherapy dosimetry, but the meeting also included one session devoted to dosimetry in diagnostic radiology. Refs, fig and tabs

  20. Dosimetry; La dosimetrie

    Energy Technology Data Exchange (ETDEWEB)

    Le Couteulx, I.; Apretna, D.; Beaugerie, M.F. [Electricite de France (EDF), 75 - Paris (France)] [and others

    2003-07-01

    Eight articles treat the dosimetry. Two articles evaluate the radiation doses in specific cases, dosimetry of patients in radiodiagnosis, three articles are devoted to detectors (neutrons and x and gamma radiations) and a computer code to build up the dosimetry of an accident due to an external exposure. (N.C.)

  1. DRDC Ottawa working standard for biological dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Segura, T.M.; Prud' homme-Lalonde, L. [Defence Research and Development Canada, Ottawa, Ontario (Canada); Thorleifson, E. [Health Canada, Gatineau, Quebec (Canada); Lachapelle, S.; Mullins, D. [JERA Consulting (Canada); Qutob, S. [Health Canada, Gatineau, Quebec (Canada); Wilkinson, D.

    2005-07-15

    This Standard provides quality assurance, quality control, and evaluation of the performance criteria for the purpose of accreditation of the Radiation Biology laboratory at Defence Research and Development Canada - Ottawa (DRDC Ottawa) using biological dosimetry to predict radiation exposure doses. The International Standard (ISO 19238) and the International Atomic Energy Association (IAEA) Technical Report Series No. 405 are used as guiding documents in preparation of this working document specific to the DRDC Ottawa Radiation Biology Laboratory. This Standard addresses: 1. The confidentiality of personal information, for the customer and the service laboratory; 2. The laboratory safety requirements; 3. The calibration sources and calibration dose ranges useful for establishing the reference dose-effect curves allowing the dose estimation from chromosome aberration frequency, and the minimum detection levels; 4. Transportation criteria for shipping of test samples to the laboratory; 5. Preparation of samples for analysis; 6. The scoring procedure for unstable chromosome aberrations used for biological dosimetry; 7. The criteria for converting a measured aberration frequency into an estimate of absorbed dose; 8. The reporting of results; 9. The quality assurance and quality control plan for the laboratory; and 10. Informative annexes containing examples of a questionnaire, instructions for customers, a data sheet for recording aberrations, a sample report and other supportive documents. (author)

  2. DRDC Ottawa working standard for biological dosimetry

    International Nuclear Information System (INIS)

    This Standard provides quality assurance, quality control, and evaluation of the performance criteria for the purpose of accreditation of the Radiation Biology laboratory at Defence Research and Development Canada - Ottawa (DRDC Ottawa) using biological dosimetry to predict radiation exposure doses. The International Standard (ISO 19238) and the International Atomic Energy Association (IAEA) Technical Report Series No. 405 are used as guiding documents in preparation of this working document specific to the DRDC Ottawa Radiation Biology Laboratory. This Standard addresses: 1. The confidentiality of personal information, for the customer and the service laboratory; 2. The laboratory safety requirements; 3. The calibration sources and calibration dose ranges useful for establishing the reference dose-effect curves allowing the dose estimation from chromosome aberration frequency, and the minimum detection levels; 4. Transportation criteria for shipping of test samples to the laboratory; 5. Preparation of samples for analysis; 6. The scoring procedure for unstable chromosome aberrations used for biological dosimetry; 7. The criteria for converting a measured aberration frequency into an estimate of absorbed dose; 8. The reporting of results; 9. The quality assurance and quality control plan for the laboratory; and 10. Informative annexes containing examples of a questionnaire, instructions for customers, a data sheet for recording aberrations, a sample report and other supportive documents. (author)

  3. Bilateral comparison 2012 between calibration laboratories dosimetry spaniards for qualities ISO: S-Cs and S-Co

    International Nuclear Information System (INIS)

    This paper describes the Protocol and the results of the comparison bilateral organized during 2012, between two Spanish calibration laboratories. The main objective of the comparison was to contribute to verification and documentation of the consistency of calibration services offered at national level, with regard to the metrological traceability of calibrations and dose levels of protection assignments radiation to gamma radiation. (Author)

  4. Accreditation - ISO/IEC 17025

    Science.gov (United States)

    Kaus, Rüdiger

    This chapter gives the background on the accreditation of testing and calibration laboratories according to ISO/IEC 17025 and sets out the requirements of this international standard. ISO 15189 describes similar requirements especially tailored for medical laboratories. Because of these similarities ISO 15189 is not separately mentioned throughout this lecture.

  5. Quality assurance in personal dosimetry of external radiation: present situation and future needs

    Energy Technology Data Exchange (ETDEWEB)

    Ma, N. [Malaysian Institue For Nuclear Technology Research (MINT), Bangi, Selangor (Malaysia)

    2006-07-01

    Whole body personal dosimetry is well established for the individual monitoring of radiation workers. High quality radiation dosimetry is essential for workers who rely upon personal dosemeters to record the amount of radiation to which they are exposed. The mandate has been given to the Personal Dosimetry, (secondary standard dosimetry laboratories) S.S.D.L., (Malaysian institute for nuclear energy research) M.I.N.T. to assure the individual monitoring for radiation workers in Malaysia. In 2005, the S.S.D.L;-M.I.N.T. supply, process and read out of personal dosemeters of nearly 13,000 dosimeters monthly, whereby. 12,000 are films and 1,000 are T.L.D.s. The objective of individual monitoring is not limited to the measurement of doses delivered to individuals, but it should demonstrate that limits of exposure have not been exceeded and that working conditions have not unexpectedly deteriorated. Dosimetry measurements are an important component of radiation protection programs and must be of high quality. The exposure of workers to radiation must be controlled and monitored in order to comply with regulatory requirements. S.S.D.L.-M.I.N.T; demonstrates that its performance is at an acceptable level by implementing overall system performance, as evidenced by the ISO 9001 certification of the Personal Dosimetry Service in 2002 and ISO/I.E.C. 17025 accreditation to the calibration laboratory in 2004. The certification and accreditation processes achieved the goal by formalizing the recognition of satisfactory performance, and providing evidence of this performance. Overall performances are assessed, personnel operating the system will be trained and are well qualified and all actions will be documented. The paper describes the overview of the Q.M.S. carried out at the S.S. D.L.-M.I.N.T.. During the implementation of Q.M.S. a few areas has been identified for future consideration. These include performance specification and type testing of dosemeters, which provide a

  6. (Re)implantation of quality system of LCR (Laboratory for Radiation Sciences) for accreditation in the standard ABNT NBR ISO/IEC 17025:2005; (Re)implantacao do sistema da qualidade do LCR para acreditacao na ABNT NBR ISO/IEC 17025:2005

    Energy Technology Data Exchange (ETDEWEB)

    Leite, Sandro P., E-mail: leite_sp@ig.com.br [Rede Sibratec, Sao Paulo, SP (Brazil); Fernandes, Elisabeth O.; David, Mariano G.; Pires, Evandro J.; Alves, Carlos F.E.; Almeida, Carlos E. [Universidade do Estado do Rio de Janeiro (UERJ), RJ (Brazil)

    2014-07-01

    This paper presents preparing procedure of the metrology laboratory (LABMETRO), which belongs Laboratorio de Ciencias Radiologicas of Rio de Janeiro , for postulating accreditation of its services metrology to INMETRO. This process, supported by the Technological Services Network SIBRATEC/FINEP for Radiation Protection and Dosimetry Technological Services, had as one of its aims to avoid possible technical barriers to the purchase services in the area of ionizing radiation laboratories. Accreditation will also enable the integration of services such laboratories in Brazilian Calibration Network (RBC). (author)

  7. Comparison of radiotherapy dosimetry for 3D-CRT, IMRT, and SBRT based on electron density calibration

    Science.gov (United States)

    Kartutik, K.; Wibowo, W. E.; Pawiro, S. A.

    2016-03-01

    Accurate calculation of dose distribution affected by inhomogeneity tissue is required in radiotherapy planning. This study was performed to determine the ratio between radiotherapy planning using 3D-CRT, IMRT, and SBRT based on a calibrated curve of CT-number in the lung for different target's shape in 3D-CRT, IMRT, and spinal cord for SBRT. Calibration curves of CT-number were generated under measurement basis and introduced into TPS, then planning was performed for 3D-CRT, IMRT, and SBRT with 7, and 15 radiation fields. Afterwards, planning evaluation was performed by comparing the DVH curve, HI, and CI. 3D-CRT and IMRT produced the lowest HI at calibration curve of CIRS 002LFC with the value 0.24 and 10. Whereas SBRT produced the lowest HI on a linear calibration curve with a value of 0.361. The highest CI in IMRT and SBRT technique achieved using a linear calibration curve was 0.97 and 1.77 respectively. For 3D-CRT, the highest CI was obtained by using calibration curve of CIRS 062M with the value of 0.45. From the results of CI and HI, it is concluded that the calibration curve of CT-number does not significantly differ with Schneider's calibrated curve, and inverse planning gives a better result than forward planning.

  8. Accredited Birth Centers

    Science.gov (United States)

    ... 717-933-9743 Accredited since January 2016 100 Bright Eyes Midwifery and Wild Rivers Women's Health Accredited ... Birthing Center-Cedar Park Accredited 1130 Cottonwood Creek Trail Building D Suite 2 Cedar Park, TX 78613 ...

  9. Evaluation of Multiple-Sampling Function used with a Microtek flatbed scanner for Radiation Dosimetry Calibration of EBT2 Film

    Science.gov (United States)

    Chang, Liyun; Ho, Sheng-Yow; Ding, Hueisch-Jy; Hwang, Ing-Ming; Chen, Pang-Yu; Lee, Tsair-Fwu

    2016-10-01

    The radiochromic EBT2 film is a widely used quality assurance device for radiation therapy. This study evaluated the film calibration performance of the multiple-sampling function, a function of the ScanWizard Pro scanning software provided by the manufacturer, when used with Microtek 9800XL plus (9800XL+) flatbed scanner. By using the PDD method, each one of the eight EBT2 films, four delivered by 290 monitor unit (MU) and four by 88 MU via 6-MV photon beams, was tightly sandwiched in a 303-cm3 water equivalent polystyrene phantom prior to irradiation. Before and after irradiation, all films were scanned using the Microtek 9800XL+ scanner with five different modes of the multiple-sampling function, which could generate the image with the averaged result of multiple-sampling. The net optical densities (netOD) on the beam central axis of film were assigned to corresponding depth doses for calibration. For each sampling mode with either delivered MU, the depth-dose uncertainty of a single film from repeated scans and that of a single scan of the four films were analyzed. Finally, the calibration error and the combined calibration uncertainty between film determined depth-doses and delivered depth-doses were calculated and evaluated for each sampling mode. All standard deviations and the calibration error were demonstrated to be unrelated to the number of sampling lines. The calibration error of the 2-line and 16-line mode was within 3 cGy and better than that of the other modes. The combined uncertainty of the 2-line mode was the lowest, which was generally less than 6 cGy except for the delivered dose around 100 cGy. The evaluation described herein revealed that the EBT2 film calibrated with the 2-line mode has relatively lower error, scanning time and combined uncertianty. Therefore, it is recommended for routine EBT2 film calibration and verification of treatment plans.

  10. Characterization of calibration curves and energy dependence GafChromic{sup TM} XR-QA2 model based radiochromic film dosimetry system

    Energy Technology Data Exchange (ETDEWEB)

    Tomic, Nada, E-mail: ntomic@roc.jgh.mcgill.ca; Quintero, Chrystian; Aldelaijan, Saad; Bekerat, Hamed; Liang, LiHeng; DeBlois, François; Devic, Slobodan [Medical Physics Unit, McGill University, Montréal, Québec H3G 1A4, Canada and Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montréal, Québec H3T 1E2 (Canada); Whiting, Bruce R. [Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania 15213 (United States); Seuntjens, Jan [Medical Physics Unit, McGill University, Montréal, Québec H3G 1A4 (Canada)

    2014-06-15

    Purpose: The authors investigated the energy response of XR-QA2 GafChromic{sup TM} film over a broad energy range used in diagnostic radiology examinations. The authors also made an assessment of the most suitable functions for both reference and relative dose measurements. Methods: Pieces of XR-QA2 film were irradiated to nine different values of air kerma in air, following reference calibration of a number of beam qualities ranging in HVLs from 0.16 to 8.25 mm Al, which corresponds to effective energy range from 12.7 keV to 56.3 keV. For each beam quality, the authors tested three functional forms (rational, linear exponential, and power) to assess the most suitable function by fitting the delivered air kerma in air as a function of film response in terms of reflectance change. The authors also introduced and tested a new parameterχ = netΔR·e{sup m} {sup netΔR} that linearizes the inherently nonlinear response of the film. Results: The authors have found that in the energy range investigated, the response of the XR-QA2 based radiochromic film dosimetry system ranges from 0.222 to 0.420 in terms of netΔR at K{sub air}{sup air} = 8 cGy. For beam qualities commonly used in CT scanners (4.03–8.25 mm Al), the variation in film response (netΔR at K{sub air}{sup air} = 8 cGy) amounts to ± 5%, while variation in K{sub air}{sup air} amounts to ± 14%. Conclusions: Results of our investigation revealed that the use of XR-QA2 GafChromic{sup TM} film is accompanied by a rather pronounced energy dependent response for beam qualities used for x-ray based diagnostic imaging purposes. The authors also found that the most appropriate function for the reference radiochromic film dosimetry would be the power function, while for the relative dosimetry one may use the exponential response function that can be easily linearized.

  11. Absolute calibration of the Gamma Knife{sup ®} Perfexion™ and delivered dose verification using EPR/alanine dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Hornbeck, Amaury, E-mail: amauryhornbeck@gmail.com, E-mail: tristan.garcia@cea.fr; Garcia, Tristan, E-mail: amauryhornbeck@gmail.com, E-mail: tristan.garcia@cea.fr [CEA, LIST, Laboratoire National Henri Becquerel, 91191 Gif-sur-Yvette Cedex (France); Cuttat, Marguerite; Jenny, Catherine [Radiotherapy Department, Medical Physics Unit, University Hospital Pitié-Salpêtrière, 75013 Paris (France)

    2014-06-15

    Purpose: Elekta Leksell Gamma Knife{sup ®} (LGK) is a radiotherapy beam machine whose features are not compliant with the international calibration protocols for radiotherapy. In this scope, the Laboratoire National Henri Becquerel and the Pitié-Salpêtrière Hospital decided to conceive a new LKG dose calibration method and to compare it with the currently used one. Furthermore, the accuracy of the dose delivered by the LGK machine was checked using an “end-to-end” test. This study also aims to compare doses delivered by the two latest software versions of the Gammaplan treatment planning system (TPS). Methods: The dosimetric method chosen is the electron paramagnetic resonance (EPR) of alanine. Dose rate (calibration) verification was done without TPS using a spherical phantom. Absolute calibration was done with factors calculated by Monte Carlo simulation (MCNP-X). For “end-to-end” test, irradiations in an anthropomorphic head phantom, close to real treatment conditions, are done using the TPS in order to verify the delivered dose. Results: The comparison of the currently used calibration method with the new one revealed a deviation of +0.8% between the dose rates measured by ion chamber and EPR/alanine. For simple fields configuration (less than 16 mm diameter), the “end-to-end” tests showed out average deviations of −1.7% and −0.9% between the measured dose and the calculated dose by Gammaplan v9 and v10, respectively. Conclusions: This paper shows there is a good agreement between the new calibration method and the currently used one. There is also a good agreement between the calculated and delivered doses especially for Gammaplan v10.

  12. Estimation of absorbed dose in clinical radiotherapy linear accelerator beams: effect of ion chamber calibration and long-term stability

    International Nuclear Information System (INIS)

    The measured dose in water at reference point in phantom is a primary parameter for planning the treatment monitor units (MU); both in conventional and intensity modulated/image guided treatments. Traceability of dose accuracy therefore still depends mainly on the calibration factor of the ion chamber/dosimeter provided by the accredited Secondary Standard Dosimetry Laboratories (SSDLs), under International Atomic Energy Agency (IAEA) network of laboratories. The data related to Nd,water calibrations, thermoluminescent dosimetry (TLD) postal dose validation, inter-comparison of different dosimeter/electrometers, and validity of Nd,water calibrations obtained from different calibration laboratories were analyzed to find out the extent of accuracy achievable. Nd,w factors in Gray/Coulomb calibrated at IBA, GmBH, Germany showed a mean variation of about 0.2% increase per year in three Farmer chambers, in three subsequent calibrations. Another ion chamber calibrated in different accredited laboratory (PTW, Germany) showed consistent Nd,w for 9 years period. The Strontium-90 beta check source response indicated long-term stability of the ion chambers within 1% for three chambers. Results of IAEA postal TL 'dose intercomparison' for three photon beams, 6 MV (two) and 15 MV (one), agreed well within our reported doses, with mean deviation of 0.03% (SD 0.87%) (n = 9). All the chamber/electrometer calibrated by a single SSDL realized absorbed doses in water within 0.13% standard deviations. However, about 1-2% differences in absorbed dose estimates observed when dosimeters calibrated from different calibration laboratories are compared in solid phantoms. Our data therefore imply that the dosimetry level maintained for clinical use of linear accelerator photon beams are within recommended levels of accuracy and uncertainities are within reported values. (author)

  13. Testing and linearity calibration of films of phenol compounds exposed to thermal neutron field for EPR dosimetry.

    Science.gov (United States)

    Gallo, S; Panzeca, S; Longo, A; Altieri, S; Bentivoglio, A; Dondi, D; Marconi, R P; Protti, N; Zeffiro, A; Marrale, M

    2015-12-01

    This paper reports the preliminary results obtained by Electron Paramagnetic Resonance (EPR) measurements on films of IRGANOX® 1076 phenols with and without low content (5% by weight) of gadolinium oxide (Gd2O3) exposed in the thermal column of the Triga Mark II reactor of LENA (Laboratorio Energia Nucleare Applicata) of Pavia (Italy). Thanks to their size, the phenolic films here presented are good devices for the dosimetry of beams with high dose gradient and which require accurate knowledge of the precise dose delivered. The dependence of EPR signal as function of neutron dose was investigated in the fluence range between 10(11) cm(-2) and 10(14) cm(-2). Linearity of EPR response was found and the signal was compared with that of commercial alanine films. Our analysis showed that gadolinium oxide (5% by weight) can enhance the thermal neutron sensitivity more than 18 times. Irradiated dosimetric films of phenolic compound exhibited EPR signal fading of about 4% after 10 days from irradiation. PMID:26242561

  14. Proceedings of the second conference on radiation protection and dosimetry

    International Nuclear Information System (INIS)

    The Second Conference on Radiation Protection and Dosimetry was held during October 31--November 3, 1988, at the Holiday Inn, Crowne Plaza Hotel in Orlando, Florida. This meeting was designed with the objectives of promoting communication among applied, research, regulatory, and standards personnel involved in radiation protection and providing them with sufficient information to evaluate their programs. To facilitate meeting these objectives, a technical program consisting of more than 75 invited and contributed oral presentations encompassing all aspects of radiation protection was prepared. General topics considered in the technical sessions included external dosimetry, internal dosimetry, calibration, standards and regulations, instrumentation, accreditation and test programs, research advances, and applied program experience. In addition, special sessions were held to afford attendees the opportunity to make short presentations of recent work or to discuss topics of general interest. This document provides a summary of the conference technical program and a partial collection of full papers for the oral presentations in order of delivery. Individual papers were processed separately for the data base

  15. Proceedings of the second conference on radiation protection and dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Swaja, R. E.; Sims, C. S. [eds.

    1988-11-01

    The Second Conference on Radiation Protection and Dosimetry was held during October 31--November 3, 1988, at the Holiday Inn, Crowne Plaza Hotel in Orlando, Florida. This meeting was designed with the objectives of promoting communication among applied, research, regulatory, and standards personnel involved in radiation protection and providing them with sufficient information to evaluate their programs. To facilitate meeting these objectives, a technical program consisting of more than 75 invited and contributed oral presentations encompassing all aspects of radiation protection was prepared. General topics considered in the technical sessions included external dosimetry, internal dosimetry, calibration, standards and regulations, instrumentation, accreditation and test programs, research advances, and applied program experience. In addition, special sessions were held to afford attendees the opportunity to make short presentations of recent work or to discuss topics of general interest. This document provides a summary of the conference technical program and a partial collection of full papers for the oral presentations in order of delivery. Individual papers were processed separately for the data base.

  16. Sixth symposium on neutron dosimetry

    International Nuclear Information System (INIS)

    This booklet contains all abstracts of papers presented in 13 sessions. Main topics: Cross sections and Kerma factors; analytical radiobiology; detectors for personnel monitoring; secondary charged particles and microdosimetric basis of q-value for neutrons; personnel dosimetry; concepts for radiation protection; ambient monitoring; TEPC and ion chambers in radiation protection; beam dosimetry; track detectors (CR-39); dosimetry at biomedical irradiation facilities; health physics at therapy facilities; calibration for radiation protection; devices for beam dosimetry (TLD and miscellaneous); therapy and biomedical irradiation facilities; treatment planning. (HP)

  17. NVLAP calibration laboratory program

    Energy Technology Data Exchange (ETDEWEB)

    Cigler, J.L.

    1993-12-31

    This paper presents an overview of the progress up to April 1993 in the development of the Calibration Laboratories Accreditation Program within the framework of the National Voluntary Laboratory Accreditation Program (NVLAP) at the National Institute of Standards and Technology (NIST).

  18. Activities developed by the biological dosimetry laboratory of the Autoridad Regulatoria Nuclear - ARN of Argentina

    International Nuclear Information System (INIS)

    Biological dosimetry (DB) allows to estimate doses absorbed in individuals exposed to ionizing radiation through the quantification of stable and unstable chromosome aberrations (SCA and UCA). The frequency of these aberrations is referred to a calibration dose response curve (in vitro) to determine the doses of the individual to the whole body. The DB is a necessary support for programs of national radiation protection and response systems in nuclear or radiological emergencies in the event of accidental or incidental, single overexposure or large scale. In this context the Laboratory of Dosimetry Biological (LDB) of the Authority Regulatory Nuclear (ARN) Argentina develops and applies different dosimeters cytogenetic from four decades ago. These dosimeters provide a fact more within the whole of the information necessary for an accidental, complementing the physical and clinical dosimetry exposure assessment. The most widely used in the DB biodosimetric method is the quantification of SCA (dicentrics and rings Central) from a sample of venous blood. The LDB is accredited for the trial, under rules IRAM 301: 2005 (ISO / IEC 17025: 2005) and ISO 19238:2004. Test applies to the immediate dosimetry evaluation of acute exposures, all or a large part of the body in the range 0,1-5 Gy. In this context the LDB is part of the Latin American network of DB (LBDNet), BioDoseNet-who and response system in radiological emergencies and nuclear IAEA-RANET, being enabled to summon the LBDNet if necessary

  19. Dosimetry Service

    CERN Multimedia

    2006-01-01

    Cern Staff and Users can now consult their dose records for an individual or an organizational unit with HRT. Please see more information on our web page: http://cern.ch/rp-dosimetry Dosimetry Service is open every morning from 8.30 - 12.00. Closed in the afternoons. We would like to remind you that dosimeters cannot be sent to customers by internal mail. Short-term dosimeters (VCT's) must always be returned to the Service after the use and must not be left on the racks in the experimental areas or in the secretariats. Dosimetry Service Tel. 7 2155 Dosimetry.service@cern.ch http://cern.ch/rp-dosimetry

  20. Evaluation of uncertainty in dosimetry of irradiator system

    International Nuclear Information System (INIS)

    This paper describes the study of uncertainties in the estimates of dosimetry irradiator system STS 0B85 of LCI IPEN/CNEN-SP. This study is relevant for determination of best measurement capability when the laboratory performs routine calibrations of measuring radiation next the optimal measures designed to radioprotection. It is also a requirement for obtaining the accreditation of the laboratory by the INMETRO. For this dosimetry was used a reference system of the laboratory composed of a electrometer and a spherical ionization chamber of 1 liter. Measurements were made at five distances selected so to include the whole range of the optical bench tests and using three attenuators filters so as to extend the measurement capability. The magnitude used for evaluation was the rate of air kerma for 137Cs and 60Co beams. Were carried out four series of measurements. It was verified the inverse square law to these series and their sets of uncertainty. Unfiltered, with one and two filters series showed good agreement with the inverse square low and the maximum uncertainty obtained was approximately 1.7%. In series with all the filters was a major deviation of the inverse square law and wide increase in uncertainty to measurements at the end of the optical bench

  1. Quality control at the Regional Centre of Nuclear Sciences chemical dosimetry laboratory

    Energy Technology Data Exchange (ETDEWEB)

    Souza, Vivianne L.B. de; Melo, Roberto T. de; Silva, Danubia B. da; Pedroza, Eryka H.; Rodrigues, Kelia R.G.; Cunha, Manuela S. da; Figueiredo, Marcela D.C. de [Centro Regional de Ciencias Nucleares (CRCN-NE/CNEN-PE), Recife, PE (Brazil); Oliveira, Aristides, E-mail: vlsouza@cnen.gov.b, E-mail: rtmelo@cnen.gov.b [Hospital de Cancer de Pernambuco, Recife, PE (Brazil)

    2011-07-01

    Standards for accreditation of laboratories as in ISO 9001 in section: 4.11 require inspection, measuring and equipment testing; likewise, IEC 17025: 2005 in section: 5.5.2 requires the equipment to be calibrated or verified before being put into use. In our laboratory, quality control is often accomplished by standards set done by the laboratory scientists themselves; however, at present, Hellma secondary calibration standards (4026 - Holmium oxide - Filters: F0, F2, F3, F4 and filter didymium - F7) have been used in order to verify if errors in the laboratory have been close to the 1-2% margin. Control graphs were made by using the results of synthetically prepared standards and standardized spectral calibration certificates. The set of secondary calibration standards permits to check the accuracy of the spectrophotometers used in research for both the absorbance in the visible spectrum (at 440, 465, 546, 590 and 635 nm wavelengths) and for the wavelengths (270, 280, 300, 320 nm) of the ultraviolet light. Filters (F0, F2, F3, F4 and F7) are stable and do not suffer the influence of temperature (the influence is negligible), the F0 filter was being used as a blank. The purpose is to verify whether the spectrometer needs adjustments, an important procedure to check absorbance stability, baseline flatness, slit width accuracy and stray radiation. The calibration tests are performed annually in our laboratory and recalibration of Hellma secondary standards is recommended every two years. The results show that the Chemical Dosimetry Laboratory in CRCN has a calibrated spectrophotometer and their synthetic standards for Fricke dosimetry could be used as an alternative method for testing the proficiency and competence of calibration laboratories in accordance with the regulations and standards. (author)

  2. [Laboratory accreditation and proficiency testing].

    Science.gov (United States)

    Kuwa, Katsuhiko

    2003-05-01

    ISO/TC 212 covering clinical laboratory testing and in vitro diagnostic test systems will issue the international standard for medical laboratory quality and competence requirements, ISO 15189. This standard is based on the ISO/IEC 17025, general requirements for competence of testing and calibration laboratories and ISO 9001, quality management systems-requirements. Clinical laboratory services are essential to patient care and therefore should be available to meet the needs of all patients and clinical personnel responsible for human health care. If a laboratory seeks accreditation, it should select an accreditation body that operates according to this international standard and in a manner which takes into account the particular requirements of clinical laboratories. Proficiency testing should be available to evaluate the calibration laboratories and reference measurement laboratories in clinical medicine. Reference measurement procedures should be of precise and the analytical principle of measurement applied should ensure reliability. We should be prepared to establish a quality management system and proficiency testing in clinical laboratories. PMID:12806918

  3. ESR Dosimetry

    International Nuclear Information System (INIS)

    ESR dosimetry is widely used for several applications such as dose assessment in accidents, medical applications and sterilization of food and other materials. In this work the dosimetric properties of natural and synthetic Hydroxyapatite, Alanine, and 2-Methylalanine are presented. Recent results on the use of a K-Band (24 GHz) ESR spectrometer in dosimetry are also presented

  4. Dosimetry Service

    CERN Multimedia

    2005-01-01

    Please remember to read your dosimeter at least once a month. Regular read-outs are vital to ensure that your personal dose is periodically monitored. Dosimeters should be read even if you have not visited the controlled areas. Dosimetry Service - Tel. 72155 http://cern.ch/rp-dosimetry

  5. Dosimetry Service

    CERN Multimedia

    2005-01-01

    Please remember to read your dosimeter at least once a month. Regular read-outs are vital to ensure that your personal dose is periodically monitored. Dosimeters should be read even if you have not visited the controlled areas. Dosimetry Service Tel. 7 2155 http://cern.ch/rp-dosimetry

  6. Dosimetry Service

    CERN Multimedia

    2005-01-01

    Please remember to read your dosimeter at least once a month. Regular read-outs are vital to ensure that your personal dose is periodically monitored. Dosimeters should be read even if you have not visited the controlled areas. Dosimetry Service - Tel. 7 2155 http://cern.ch/rp-dosimetry

  7. Computational dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Siebert, B.R.L.; Thomas, R.H.

    1996-01-01

    The paper presents a definition of the term ``Computational Dosimetry`` that is interpreted as the sub-discipline of computational physics which is devoted to radiation metrology. It is shown that computational dosimetry is more than a mere collection of computational methods. Computational simulations directed at basic understanding and modelling are important tools provided by computational dosimetry, while another very important application is the support that it can give to the design, optimization and analysis of experiments. However, the primary task of computational dosimetry is to reduce the variance in the determination of absorbed dose (and its related quantities), for example in the disciplines of radiological protection and radiation therapy. In this paper emphasis is given to the discussion of potential pitfalls in the applications of computational dosimetry and recommendations are given for their avoidance. The need for comparison of calculated and experimental data whenever possible is strongly stressed.

  8. Training Accreditation Program

    International Nuclear Information System (INIS)

    The Training Accreditation Program establishes the objectives and criteria against which DOE nuclear facility training is evaluated to determine its readiness for accreditation. Training programs are evaluated against the accreditation objectives and criteria by facility personnel during the initial self-evaluation process. From this self-evaluation, action plans are made by the contractor to address the scope of work necessary in order to upgrade any deficiencies noted. This scope of work must be formally documented in the Training Program Accreditation Plan. When reviewed and approved by the responsible Head of the Field Organization and cognizant Program Secretarial Office, EH-1 concurrence is obtained. This plan then becomes the document which guides accreditation efforts for the contractor

  9. Dosimetry in diagnostic radiology

    International Nuclear Information System (INIS)

    Dosimetry is an area of increasing importance in diagnostic radiology. There is a realisation amongst health professionals that the radiation dose received by patients from modern X-ray examinations and procedures can be at a level of significance for the induction of cancer across a population, and in some unfortunate instances, in the acute damage to particular body organs such as skin and eyes. The formulation and measurement procedures for diagnostic radiology dosimetry have recently been standardised through an international code of practice which describes the methodologies necessary to address the diverging imaging modalities used in diagnostic radiology. Common to all dosimetry methodologies is the measurement of the air kerma from the X-ray device under defined conditions. To ensure the accuracy of the dosimetric determination, such measurements need to be made with appropriate instrumentation that has a calibration that is traceable to a standards laboratory. Dosimetric methods are used in radiology departments for a variety of purposes including the determination of patient dose levels to allow examinations to be optimized and to assist in decisions on the justification of examination choices. Patient dosimetry is important for special cases such as for X-ray examinations of children and pregnant patients. It is also a key component of the quality control of X-ray equipment and procedures.

  10. Dosimetry in diagnostic radiology.

    Science.gov (United States)

    Meghzifene, Ahmed; Dance, David R; McLean, Donald; Kramer, Hans-Michael

    2010-10-01

    Dosimetry is an area of increasing importance in diagnostic radiology. There is a realisation amongst health professionals that the radiation dose received by patients from modern X-ray examinations and procedures can be at a level of significance for the induction of cancer across a population, and in some unfortunate instances, in the acute damage to particular body organs such as skin and eyes. The formulation and measurement procedures for diagnostic radiology dosimetry have recently been standardised through an international code of practice which describes the methodologies necessary to address the diverging imaging modalities used in diagnostic radiology. Common to all dosimetry methodologies is the measurement of the air kerma from the X-ray device under defined conditions. To ensure the accuracy of the dosimetric determination, such measurements need to be made with appropriate instrumentation that has a calibration that is traceable to a standards laboratory. Dosimetric methods are used in radiology departments for a variety of purposes including the determination of patient dose levels to allow examinations to be optimized and to assist in decisions on the justification of examination choices. Patient dosimetry is important for special cases such as for X-ray examinations of children and pregnant patients. It is also a key component of the quality control of X-ray equipment and procedures. PMID:20655679

  11. Third conference on radiation protection and dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    1991-01-01

    This conference has been designed with the objectives of promoting communication among applied, research, regulatory, and standards personnel involved in radiation protection and providing them with sufficient information to evaluate their programs. To partly fulfill these objectives, a technical program consisting of more than 75 invited and contributed oral presentations encompassing all aspects of radiation protection has been prepared. General topics include external dosimetry, internal dosimetry, instruments, regulations and standards, accreditation and test programs, research advances, and applied program experience. This publication provides a summary of the technical program and a collection of abstracts of the oral presentations.

  12. Third conference on radiation protection and dosimetry

    International Nuclear Information System (INIS)

    This conference has been designed with the objectives of promoting communication among applied, research, regulatory, and standards personnel involved in radiation protection and providing them with sufficient information to evaluate their programs. To partly fulfill these objectives, a technical program consisting of more than 75 invited and contributed oral presentations encompassing all aspects of radiation protection has been prepared. General topics include external dosimetry, internal dosimetry, instruments, regulations and standards, accreditation and test programs, research advances, and applied program experience. This publication provides a summary of the technical program and a collection of abstracts of the oral presentations

  13. [Accreditation of forensic laboratories].

    Science.gov (United States)

    Sołtyszewski, Ireneusz

    2010-01-01

    According to the framework decision of the European Union Council, genetic laboratories which perform tests for the benefit of the law enforcement agencies and the administration of justice are required to obtain a certificate of accreditation testifying to compliance with the PN EN ISO/IEC 17025:2005 standard. The certificate is the official confirmation of the competence to perform research, an acknowledgement of credibility, impartiality and professional independence. It is also the proof of establishment, implementation and maintenance of an appropriate management system. The article presents the legal basis for accreditation, the procedure of obtaining the certificate of accreditation and selected elements of the management system. PMID:21863740

  14. Accreditation of nuclear engineering programs

    International Nuclear Information System (INIS)

    The American Nuclear Society (ANS) Professional Development and Accreditation Committee (PDAC) has the responsibility for accreditation of engineering and technology programs for nuclear and similarly named programs. This committee provides society liaison with the Accreditation Board for Engineering and Technology (ABET), is responsible for the appointment and training of accreditation visitors, nomination of members for the ABET Board and Accreditation Commissions, and review of the criteria for accreditation of nuclear-related programs. The committee is composed of 21 members representing academia and industry. The ABET consists of 19 participating bodies, primarily professional societies, and 4 affiliate bodies. Representation on ABET is determined by the size of the professional society and the number of programs accredited. The ANS, as a participating body, has one member on the ABET board, two members on the Engineering Accreditation Commission, and one on the Technology Accreditation Commission. The ABET board sets ABET policy and the commissions are responsible for accreditation visits

  15. SU-E-I-24: Design and Fabrication of a Multi-Functional Neck and Thyroid Phantom for Medical Dosimetry and Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Mehdizadeh, S; Sina, S [Radiation Research Center, Shiraz University, Shiraz (Iran, Islamic Republic of); Karimipourfard, M; Lotfalizadeh, F [Nuclear Engineering department, Shiraz University, Shiraz (Iran, Islamic Republic of); Faghihi, R [Radiation Research Center, Shiraz University, Shiraz (Iran, Islamic Republic of); Nuclear Engineering department, Shiraz University, Shiraz (Iran, Islamic Republic of); Babaei, A [Shiraz University of medical sciences, Shiraz (Iran, Islamic Republic of)

    2014-06-01

    Purpose: The purpose of this study is the design and fabrication of a multipurpose anthropomorphic neck and thyroid phantom for use in medical applications (i.e. quality control of images in nuclear medicine, and dosimetry). Methods: The designed neck phantom is composed of seven elliptic cylindrical slices with semi-major axis of 14 and semi-minor axis of 12.5 cm, each having the thickness of 2cm. Thyroid gland, bony part of the neck, and the wind pipe were also built inside the neck phantom. Results: The phantom contains some removable plugs,inside and at its surface to accommodate the TLD chips with different shapes and dimensions, (i.e. rod, cylindrical and cubical TLD chips)for the purpose of medical dosimetry (i.e. in radiology, radiotherapy, and nuclear medicine). For the purpose of quality control of images in nuclear medicine, the removable thyroid gland was built to accommodate the radioactive iodine. The female and male thyroid glands were built in two sizes separately. Conclusion: The designed phantom is a multi-functional phantom which is applicable for dosimetry in diagnostic radiology, radiotherapy, and quality control of images in nuclear medicine.

  16. Dosimetry Service

    CERN Multimedia

    Dosimetry Service

    2005-01-01

    Please remember to read your dosimeter at least once every month. A regular read-out is indispensable to ensure periodic monitoring of your personal dose. You must read your dosimeter even if you have not visited the controlled areas. Film badges are no longer valid at CERN and holders of film badges are no longer allowed to enter the controlled radiation areas or work with a source. Dosimetry Service Tel. 72155 http://cern.ch/rp-dosimetry

  17. Dosimetry methods

    DEFF Research Database (Denmark)

    McLaughlin, W.L.; Miller, A.; Kovacs, A.;

    2003-01-01

    Chemical and physical radiation dosimetry methods, used for the measurement of absorbed dose mainly during the practical use of ionizing radiation, are discussed with respect to their characteristics and fields of application.......Chemical and physical radiation dosimetry methods, used for the measurement of absorbed dose mainly during the practical use of ionizing radiation, are discussed with respect to their characteristics and fields of application....

  18. Secondary calibration laboratory for dosimetry in levels of therapy at the University of Santiago; Laboratorio secundario de calibracion para dosimetria en niveles de terapia en la Universidad de Santiago

    Energy Technology Data Exchange (ETDEWEB)

    Gomez Rodriguez, F.; Gonzalez Castano, D. M.; Pazos Alvarez, A.

    2011-07-01

    A basic inherent benefits provided by the existence of a traceability chain radiation in any application, add the legal requirement for hospitals as pointed to by the RO. 1566/1998, which sets quality standards in radiotherapy. The decree attributed to hospital specialists radio physics in article 10 the responsibility for determining the acceptance and initial reference state of radiation generating equipment for therapeutic purposes, and the establishment and implementation of quality control programs associated and technical and physical aspects of radiation dosimetry. Different international organizations such as ICRU and IAEA recommendations on maintaining the accuracy of the dose delivered to patients in general, should be placed at least 5% considering the whole chain irradiation. In order to achieve this purpose it is necessary to establish programs of quality control and calibration dosimetric regular basis. The protocol of the IAEA TRS398 recommended dose calibration in water because it is a quantity of interest closest to clinical use and allows a relative uncertainty in the calibration environment reduced to 1%.. (Author)

  19. Performance testing of dosimetry processors, status of NRC rulemaking for improved personnel dosimetry processing, and some beta dosimetry and instrumentation problems observed by NRC regional inspectors

    International Nuclear Information System (INIS)

    Early dosimetry processor performance studies conducted between 1967 and 1979 by several different investigators indicated that a significant percentage of personnel dosimetry processors may not be performing with a reasonable degree of accuracy. Results of voluntary performance testing of US personnel dosimetry processors against the final Health Physics Society Standard, Criteria for Testing Personnel Dosimetry Performance by the University of Michigan for the Nuclear Regulatory Commission (NRC) will be summarized with emphasis on processor performance in radiation categories involving beta particles and beta particles and photon mixtures. The current status of the NRC's regulatory program for improved personnel dosimetry processing will be reviewed. The NRC is proposing amendments to its regulations, 10 CFR Part 20, that would require its licensees to utilize specified personnel dosimetry services from processors accredited by the National Voluntary Laboratory Accreditation Program of the National Bureau of Standards. Details of the development and schedule for implementation of the program will be highlighted. Finally, selected beta dosimetry and beta instrumentation problems observed by NRC Regional Staff during inspections of NRC licensed facilities will be discussed

  20. The Next Accreditation System.

    Science.gov (United States)

    Kirk, Lynne M

    2016-02-01

    The Accreditation Council for Graduate Medical Education has implemented a new accreditation system for graduate medical education in the United States. This system, called the Next Accreditation System, focuses on more continuous monitoring of the outcomes of residency training, and for high-quality programs, less on the detailed processes of that training. This allows programs to innovate to best meet the needs of their trainees and communities. This new system also reviews the clinical learning environment at each institution sponsoring graduate medical education, focusing on professionalism, trainee supervision, duty hour and fatigue management, care transitions, and integration of residents into patient safety and health care quality. This Next Accreditation System is too new to fully assess its outcomes in better preparing residents for medical practice. Assessments of its early implementation, however, suggest we can expect such outcomes in the near future. PMID:26859375

  1. [Accreditation of medical laboratories].

    Science.gov (United States)

    Horváth, Andrea Rita; Ring, Rózsa; Fehér, Miklós; Mikó, Tivadar

    2003-07-27

    In Hungary, the National Accreditation Body was established by government in 1995 as an independent, non-profit organization, and has exclusive rights to accredit, amongst others, medical laboratories. The National Accreditation Body has two Specialist Advisory Committees in the health care sector. One is the Health Care Specialist Advisory Committee that accredits certifying bodies, which deal with certification of hospitals. The other Specialist Advisory Committee for Medical Laboratories is directly involved in accrediting medical laboratory services of health care institutions. The Specialist Advisory Committee for Medical Laboratories is a multidisciplinary peer review group of experts from all disciplines of in vitro diagnostics, i.e. laboratory medicine, microbiology, histopathology and blood banking. At present, the only published International Standard applicable to laboratories is ISO/IEC 17025:1999. Work has been in progress on the official approval of the new ISO 15189 standard, specific to medical laboratories. Until the official approval of the International Standard ISO 15189, as accreditation standard, the Hungarian National Accreditation Body has decided to progress with accreditation by formulating explanatory notes to the ISO/IEC 17025:1999 document, using ISO/FDIS 15189:2000, the European EC4 criteria and CPA (UK) Ltd accreditation standards as guidelines. This harmonized guideline provides 'explanations' that facilitate the application of ISO/IEC 17025:1999 to medical laboratories, and can be used as a checklist for the verification of compliance during the onsite assessment of the laboratory. The harmonized guideline adapted the process model of ISO 9001:2000 to rearrange the main clauses of ISO/IEC 17025:1999. This rearrangement does not only make the guideline compliant with ISO 9001:2000 but also improves understanding for those working in medical laboratories, and facilitates the training and education of laboratory staff. With the

  2. Dosimetry in radiotherapy. V.2

    International Nuclear Information System (INIS)

    A series of symposia on dosimetry in medicine and biology have been held by the IAEA in co-operation with WHO. The present symposium was the first one focusing on ''Dosimetry in Radiotherapy''. The papers presented reflected the different steps in the calibration chain such as the calibration standards established by the National Standards Laboratories and the conversion of the reading of calibrated instruments to the desired quantity, i.e. absorbed dose to water at a reference point in the user's beam at the radiotherapy clinic. The programme further examined the procedures necessary for optimization of the treatment of the patient, such as treatment planning methods, dose distribution studies, new techniques of dose measurement, improvements in the physical dose distributions/conformation therapy and special problems involved in total body treatments. Results of quality assurance in radiotherapy were presented from local hospitals as well as from national and international studies. Refs, figs and tabs

  3. Dosimetry in radiotherapy. V.1

    International Nuclear Information System (INIS)

    A series of symposia on dosimetry in medicine and biology have been held by the IAEA in co-operation with WHO. The present symposium was the first one focusing on ''Dosimetry in Radiotherapy''. The papers presented reflected the different steps in the calibration chain such as the calibration standards established by the National Standards Laboratories and the conversion of the reading of calibrated instruments to the desired quantity, i.e. absorbed dose to water at a reference point in the user's beam at the radiotherapy clinic. The programme further examined the procedures necessary for optimization of the treatment of the patient, such as treatment planning methods, dose distribution studies, new techniques of dose measurement, improvements in the physical dose distributions/conformation therapy and special problems involved in total body treatments. Results of quality assurance in radiotherapy were presented from local hospitals as well as from national and international studies. Refs, figs and tabs

  4. 42 CFR 424.58 - Accreditation.

    Science.gov (United States)

    2010-10-01

    ... findings by the accreditation organization and findings by CMS or its designated survey team on standards... to designate and approve one or more independent accreditation organizations for purposes of... for accreditation organizations. (1) An independent accreditation organization applying for...

  5. 热释光剂量测量系统检定的质量保证及刻度因子的应用%Quality assurance for verification of thermoluminescence dosimetry system and application of calibration factors

    Institute of Scientific and Technical Information of China (English)

    李秀芹; 赵进沛; 米宁; 杨新芳; 侯金兵

    2012-01-01

    [ Objective] To ensure the reliability of individual dose monitoring data of radiation exposed workers, explore the influencing factors during verification process of thermoluminescence (TL) dosimetry system, calculate the linearity and the calibration factors of different energy respond, and determine the calibration factors that were suitable to different types of radiation exposed workers. [Methods] According to JJG 593-2006, two schemes were designed for verification. Scheme one; the chosen TL detectors were placed directly on the phantom and irradiated by different dosage 137 Cs--y ray radiation. Scheme two; the chosen TL detectors were worn by boxes, then placed on the phantom and irradiated by different dosage 137Cs~7 radiation and different energy X rays. The linearity and calibration factors were calculated. [ Results]There was significant difference between the detectors placed directly on the phantom and detectors worn by boxes, and the difference between calibration factors of different energy respond was significant , too. [ Conclusion] In the course of the verification of TL dosimetry system, the balance condition should be controlled strictly , and it is important to adopt the detector boxes for balance. The low energy X-ray radiation calibration factors should be used in radiation diagnosis group, and the -/-ray radiation calibration factors is suitable to radiation therapy group. The application of two calibration factors can get more accurate monitoring data.%目的 为保证辐射受照人员个人剂量监测数据的可靠性,探讨热释光剂量测量系统检定过程影响因素,计算线性和不同能响刻度因子,确定不同类别照射群体所使用的刻度因子.方法 依据标准JJG 593 - 2006,设计2种方案进行系统检定.方案1:已筛选退火后的探测器片放于有机玻璃平衡板中,分别予以不同剂量的137 Cs γ射线照射;方案2:将探测器片放于剂量计盒中,置于与方案1相同位置,予γ

  6. From Evaluation to Accreditation

    DEFF Research Database (Denmark)

    Rasmussen, Palle

    for policy. In the state controlled and public financed Danish higher education system quality assessment became institutionalised in a national agency, the "evaluation centre", which was to do recurrent assessment of all higher education programmes. This was later given up. Recently accreditation......Quality was introduced as political priority in Danish higher education during the 1980ties, associated with new public management as well as with new liberalism and conservatism. As a political goal the concept of quality has a paradoxical character because it does not lay out any definite course...... of education programmes has been introduced, also in the form of a national agency with the mission of accrediting all higher education programmes. The paper discusses reasons for and problems in this approach, and the more general social functions of quality assessment and accreditation....

  7. [Accreditation in health care].

    Science.gov (United States)

    Fügedi, Gergely; Lám, Judit; Belicza, Éva

    2016-01-24

    Besides the rapid development of healing procedures and healthcare, efficiency of care, institutional performance and safe treatment are receiving more and more attention in the 21st century. Accreditation, a scientifically proven tool for improving patient safety, has been used effectively in healthcare for nearly a hundred years, but only started to spread worldwide since the 1990s. The support and active participation of medical staff are determining factors in operating and getting accross the nationally developed, upcoming Hungarian accreditation system. However, this active assistance cannot be expected without the participants' understanding of the basic goals and features of the system. The presence of the ISO certification in Hungary, well-known by healthcare professionals, further complicates the understanding and orientation among quality management and improvement systems. This paper aims to provide an overview of the history, goals, function and importance of healthcare accreditation, and its similarities and differences regarding ISO certification. PMID:26772826

  8. (Biological dosimetry)

    Energy Technology Data Exchange (ETDEWEB)

    Preston, R.J.

    1990-12-17

    The traveler attended the 1st International Conference on Biological Dosimetry in Madrid, Spain. This conference was organized to provide information to a general audience of biologists, physicists, radiotherapists, industrial hygiene personnel and individuals from related fields on the current ability of cytogenetic analysis to provide estimates of radiation dose in cases of occupational or environmental exposure. There is a growing interest in Spain in biological dosimetry because of the increased use of radiation sources for medical and occupational uses, and with this the anticipated and actual increase in numbers of overexposure. The traveler delivered the introductory lecture on Biological Dosimetry: Mechanistic Concepts'' that was intended to provide a framework by which the more applied lectures could be interpreted in a mechanistic way. A second component of the trip was to provide advice with regard to several recent cases of overexposure that had been or were being assessed by the Radiopathology and Radiotherapy Department of the Hospital General Gregorio Maranon'' in Madrid. The traveler had provided information on several of these, and had analyzed cells from some exposed or purportedly exposed individuals. The members of the biological dosimetry group were referred to individuals at REACTS at Oak Ridge Associated Universities for advice on follow-up treatment.

  9. Is Gerontology Ready for Accreditation?

    Science.gov (United States)

    Haley, William E.; Ferraro, Kenneth F.; Montgomery, Rhonda J. V.

    2012-01-01

    The authors review widely accepted criteria for program accreditation and compare gerontology with well-established accredited fields including clinical psychology and social work. At present gerontology lacks many necessary elements for credible professional accreditation, including defined scope of practice, applied curriculum, faculty with…

  10. Accreditation and the Credit Hour.

    Science.gov (United States)

    Wellman, James V.

    2003-01-01

    Reviews the role that accreditation plays in defining and enforcing the credit-hour measure. Regional accreditation agencies are generally more flexible in terms of defining credit hours than are national agencies, which are more rigid in their expectations. Specialized accrediting agencies usually make the least mention of credit units. (SLD)

  11. Is gerontology ready for accreditation?

    Science.gov (United States)

    Haley, William E; Ferraro, Kenneth F; Montgomery, Rhonda J V

    2012-01-01

    The authors review widely accepted criteria for program accreditation and compare gerontology with well-established accredited fields including clinical psychology and social work. At present gerontology lacks many necessary elements for credible professional accreditation, including defined scope of practice, applied curriculum, faculty with applied professional credentials, and resources necessary to support professional credentialing review. Accreditation with weak requirements will be dismissed as "vanity" accreditation, and strict requirements will be impossible for many resource-poor programs to achieve, putting unaccredited programs at increased risk for elimination. Accreditation may be appropriate in the future, but it should be limited to professional or applied gerontology, perhaps for programs conferring bachelor's or master's degrees. Options other than accreditation to enhance professional skills and employability of gerontology graduates are discussed. PMID:22289064

  12. Is gerontology ready for accreditation?

    Science.gov (United States)

    Haley, William E; Ferraro, Kenneth F; Montgomery, Rhonda J V

    2012-01-01

    The authors review widely accepted criteria for program accreditation and compare gerontology with well-established accredited fields including clinical psychology and social work. At present gerontology lacks many necessary elements for credible professional accreditation, including defined scope of practice, applied curriculum, faculty with applied professional credentials, and resources necessary to support professional credentialing review. Accreditation with weak requirements will be dismissed as "vanity" accreditation, and strict requirements will be impossible for many resource-poor programs to achieve, putting unaccredited programs at increased risk for elimination. Accreditation may be appropriate in the future, but it should be limited to professional or applied gerontology, perhaps for programs conferring bachelor's or master's degrees. Options other than accreditation to enhance professional skills and employability of gerontology graduates are discussed.

  13. Dosimetry Service

    CERN Multimedia

    Dosimetry Service

    2005-01-01

    Please remember to read your dosimeter every month at least once and preferably during the first week. A regular read-out is indispensable in order to ensure a periodic monitoring of the personal dose. You should read your dosimeter even if you have not visited the controlled areas. If you still have the old dosimeter (film badge), please send it immediately for evaluation to us (Bdg 24 E-011). After January 2005 there will be no developing process for the old film system. Information for Contractors: Please remember also to bring the form ‘Confirm Reception of a CERN Dosimeter' signed with ‘Feuille d'enregistrement du CERN'. Without these forms the dosimeter cannot be assigned. Thank you for your cooperation. Dosimetry Service Tel 767 2155 http://cern.ch/rp-dosimetry

  14. Accredited dose measurements for validation of radiation sterilized products

    DEFF Research Database (Denmark)

    Miller, A.

    1993-01-01

    for control of radiation sterilization. The accredited services include: 1. 1. Irradiation of dosimeters and test samples with cobalt-60 gamma rays. 2. 2. Irradiation of dosimeters and test samples with 10 MeV electrons. 3. 3. Issue of and measurement with calibrated dosimeters. 4. 4. Measurement...

  15. Environmental dosimetry of radon-222 and daughters: measurement of absolute calibration factors of CR-39 considering the plate-out effects and environmental factors

    International Nuclear Information System (INIS)

    The subject of this work concerns with the measurement of absolute calibration factors for the use of CR-39 as an absolute detector in indoor and daughters monitoring. Up to now the usefulness of calibration factors was restricted to environmental conditions equal (or very close) to those worthing during their determinations. This fact is consequence of the difficulties related to the understanding of the plate-out properties of radon daughters activity in the air. The plate-out effects on radon daughters monitoring performed by SSNTDs are studied. Our experimental results are in agreement with those of other authors about the great sensitivity of CR-39 to the plate-out effects, fact that recommended its use in this work. Being succeeded in the employment of CR-39 as an alpha-spectrometer we concluded that some important information (like the radon daughters deposition rates on the walls of an environment) can be achieved. The knowledge about the behavior of plate-out made possible the determination of the ranges in zenithal angle and energy where CR-39 can detect alpha-particles with efficiency of 100%, at our conditions of track observation. In this way, we obtained calibration factors for CR-39 that are weakly dependent on environmental conditions. We think that these results can contribute to the improvement of RD (Radiation Detector) detection techniques. (author). 159 refs, 106 figs, 05 tabs

  16. Activities developed by the biological dosimetry laboratory of the Autoridad Regulatoria Nuclear - ARN of Argentina; Actividades desarrolladas por el laboratorio de dosimetria biologica de la Autoridad Regulatoria Nuclear de Argentina

    Energy Technology Data Exchange (ETDEWEB)

    Radl, A.; Sapienza, C.E.; Taja, M.R.; Bubniak, R.; Deminge, M.; Di Giorgio, M., E-mail: csapienza@arn.gob.ar [Autoridad Regulatoria Nuclear (ARN), Buenos Aires (Argentina)

    2013-07-01

    Biological dosimetry (DB) allows to estimate doses absorbed in individuals exposed to ionizing radiation through the quantification of stable and unstable chromosome aberrations (SCA and UCA). The frequency of these aberrations is referred to a calibration dose response curve (in vitro) to determine the doses of the individual to the whole body. The DB is a necessary support for programs of national radiation protection and response systems in nuclear or radiological emergencies in the event of accidental or incidental, single overexposure or large scale. In this context the Laboratory of Dosimetry Biological (LDB) of the Authority Regulatory Nuclear (ARN) Argentina develops and applies different dosimeters cytogenetic from four decades ago. These dosimeters provide a fact more within the whole of the information necessary for an accidental, complementing the physical and clinical dosimetry exposure assessment. The most widely used in the DB biodosimetric method is the quantification of SCA (dicentrics and rings Central) from a sample of venous blood. The LDB is accredited for the trial, under rules IRAM 301: 2005 (ISO / IEC 17025: 2005) and ISO 19238:2004. Test applies to the immediate dosimetry evaluation of acute exposures, all or a large part of the body in the range 0,1-5 Gy. In this context the LDB is part of the Latin American network of DB (LBDNet), BioDoseNet-who and response system in radiological emergencies and nuclear IAEA-RANET, being enabled to summon the LBDNet if necessary.

  17. Laboratory Accreditation in Argentina.

    Science.gov (United States)

    Acuña, María Amelia; Collino, Cesar; Chiabrando, Gustavo A

    2015-11-01

    Laboratory accreditation is an essential element in the healthcare system since it contributes substantially to decision-making, in the prevention, diagnosis, treatment and follow-up of the health status of the patients, as well as in the organization and management of public healthcare. Therefore, the clinical biochemistry professional works continuously to provide reliable results and contributes to the optimization of operational logistics and integration of a laboratory into the health system. ISO 15189 accreditation, ensures compliance of the laboratory to minimize instances of error through the planning, prevention, implementation, evaluation and improvement of its procedures, which provides skill areas that involve both training undergraduate and graduate professionals in clinical biochemistry. PMID:27683497

  18. FIS accreditation guidelines

    OpenAIRE

    Ojanen, Pinja

    2010-01-01

    The International Ski Federation (FIS) was founded to support and develop the sport of skiing 100 years ago. Since then skiing has grown in importance and has become more popular. Nowadays ski companies, athletes and family members, but also fan clubs, national ski associations, sponsors and local organizers get involved for a sporting event. The aim of this project-based thesis was to create guidelines and make the FIS accreditation system as user-friendly as possible. For the groups (ser...

  19. Mammography accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Wilcox, P.

    1993-12-31

    In the mid-1980`s, the movement toward the use of dedicated mammography equipment provided significant improvement in breast cancer detection. However, several studies demonstrated that this change was not sufficient to ensure optimal image quality at a low radiation dose. In particular, the 1985 Nationwide Evaluation of X-ray Trends identified the wide variations in image quality and radiation dose, even from dedicated units. During this time period, the American Cancer Society (ACS) launched its Breast Cancer Awareness Screening Campaign. However, there were concerns about the ability of radiology to respond to the increased demand for optimal screening examinations that would result from the ACS program. To respond to these concerns, the ACS and the American College of Radiology (ACR) established a joint committee on mammography screening in 1986. After much discussion, it was decided to use the ACR Diagnostic Practice Accreditation Program as a model for the development of a mammography accreditation program. However, some constraints were required in order to make the program meet the needs of the ACS. This voluntary, peer review program had to be timely and cost effective. It was determined that the best way to address these needs would be to conduct the program by mail. Finally, by placing emphasis on the educational nature of the program, it would provide an even greater opportunity for improving mammographic quality. The result of this effort was that, almost six years ago, in May 1987, the pilot study for the ACR Mammography Accreditation Program (MAP) began, and in August of that year, the first applications were received. In November 1987, the first 3-year accreditation certificates were awarded.

  20. Technical basis document for internal dosimetry

    International Nuclear Information System (INIS)

    This document provides the technical basis for the Chem-Nuclear Geotech (Geotech) internal dosimetry program. Geotech policy describes the intentions of the company in complying with radiation protection standards and the as low as reasonably achievable (ALARA) program. It uses this policy and applicable protection standards to derive acceptable methods and levels of bioassay to assure compliance. The models and computational methods used are described in detail within this document. FR-om these models, dose- conversion factors and derived limits are computed. These computations are then verified using existing documentation and verification information or by demonstration of the calculations used to obtain the dose-conversion factors and derived limits. Recommendations for methods of optimizing the internal dosimetry program to provide effective monitoring and dose assessment for workers are provided in the last section of this document. This document is intended to be used in establishing an accredited dosimetry program in accordance with expected Department of Energy Laboratory Accreditation Program (DOELAP) requirements for the selected radionuclides provided in this document, including uranium mill tailing mixtures. Additions and modifications to this document and procedures derived FR-om this document are expected in the future according to changes in standards and changes in programmatic mission

  1. Dosimetry standards for radiation processing

    International Nuclear Information System (INIS)

    For irradiation treatments to be reproducible in the laboratory and then in the commercial environment, and for products to have certified absorbed doses, standardized dosimetry techniques are needed. This need is being satisfied by standards being developed by experts from around the world under the auspices of Subcommittee E10.01 of the American Society for Testing and Materials (ASTM). In the time period since it was formed in 1984, the subcommittee has grown to 150 members from 43 countries, representing a broad cross-section of industry, government and university interests. With cooperation from other international organizations, it has taken the combined part-time effort of all these people more than 13 years to complete 24 dosimetry standards. Four are specifically for food irradiation or agricultural applications, but the majority apply to all forms of gamma, x-ray, Bremsstrahlung and electron beam radiation processing, including dosimetry for sterilization of health care products and the radiation processing of fruits, vegetables, meats, spices, processed foods, plastics, inks, medical wastes and paper. An additional 6 standards are under development. Most of the standards provide exact procedures for using individual dosimetry systems or for characterizing various types of irradiation facilities, but one covers the selection and calibration of dosimetry systems, and another covers the treatment of uncertainties. Together, this set of standards covers essentially all aspects of dosimetry for radiation processing. The first 20 of these standards have been adopted in their present form by the International Organization of Standardization (ISO), and will be published by ISO in 1999. (author)

  2. Calibration of CT density values in dosimetry verification of intensity modulated radiation therapy%调强适形放射治疗剂量验证中CT密度值的校准分析

    Institute of Scientific and Technical Information of China (English)

    庄名赞; 吴仁华; 邱庆春; 彭逊; 陆佳扬; 陈志坚

    2011-01-01

    Objecfive Based on intensity modulated radiation therapy (IMRT)phantom,the impact of CT-to-density conversion curve on dosimetry verification of IMRT is investigated and calibrated.Methods The electron density phantom was used to establish the CT-to-density conversion curve in radiation treatment planning system.IMRT plans of 12 nasopharynx carcinoma patients were chosen,copied to IMRT phantom and computed for the dose distribution.For each plan a measured point was put at the place where the dose was well-distributed and its dose value was measured using the ionization chamber.The physical density of IMRT phantom and its CTvalue were input into the planning system,to make a calibration for the CT-to-density conversion curve.The dose distribution was recomputed for each IMRT plan.Other parameters were kept the same in the plans and the differences between the computed dose values before and after correction were compared with the measured values.Results In 12 nasopharynx carcinoma IMRT plans,the average error of computed dose values was 1.96%±0.87% before correction and 0.63%±0.74%after correction,compared with measured values.The error between measured values and computed values after correction was less than ±2% whereas the maximum error of computed values before correction was 3.24%.Conclusions The computed dose values are closer to the measured values when using the calibrated CT-to-density conversion curve.The CT density values of IMRT phantom should be verified before usage,so as to increase the accuracy of IMRT dosimetry verification.%目的 基于调强适形放射治疗(IMRT)验证体模,分析放射治疗计划系统中CT密度转换曲线在IMRT剂量验证中的影响,并加以校准.方法 利用CT电子密度体模刻度计划系统中的CT密度转换曲线,取12例鼻咽癌患者的IMRT计划,分别移植至IMRT验证体模,计算剂量分布,并于剂量均匀处放置测量点,利用电离室测量出其实际剂量.将IMRT验证体模的物

  3. Topics in radiation dosimetry radiation dosimetry

    CERN Document Server

    1972-01-01

    Radiation Dosimetry, Supplement 1: Topics in Radiation Dosimetry covers instruments and techniques in dealing with special dosimetry problems. The book discusses thermoluminescence dosimetry in archeological dating; dosimetric applications of track etching; vacuum chambers of radiation measurement. The text also describes wall-less detectors in microdosimetry; dosimetry of low-energy X-rays; and the theory and general applicability of the gamma-ray theory of track effects to various systems. Dose equivalent determinations in neutron fields by means of moderator techniques; as well as developm

  4. Radio-analysis. Applications: biological dosimetry; Radioanalyse. Applications: dosage biologique

    Energy Technology Data Exchange (ETDEWEB)

    Bourrel, F. [CEA Saclay, INSTN, Institut National des Sciences et Techniques Nucleaires, 91 - Gif-sur-Yvette (France); Courriere, Ph. [UFR de Pharmacie, 31 - Toulouse (France)

    2003-06-01

    Radioisotopes have revolutionized the medical biology. Radio-immunology remains the reference measurement of the infinitely small in biology. Constant efforts have been performed to improve the simpleness, detectability and fastness of the method thanks to an increasing automation. This paper presents: 1 - the advantages of compounds labelling and the isotopic dilution; 2 - the antigen-antibody system: properties, determination of the affinity constant using the Scatchard method; 3 - radio-immunologic dosimetry: competitive dosimetry (radioimmunoassay), calibration curve and mathematical data processing, application to the free thyroxine dosimetry, immunoradiometric dosimetry (immunoradiometric assay), evaluation of the analytical efficiency of a radioimmunoassay; 4 - detection of the radioactive signal (solid and liquid scintillation). (J.S.)

  5. Calibration of farmer dosemeters

    International Nuclear Information System (INIS)

    The Farmer Dosemeters of Atomic Energy Medical Centre (AEMC) Jamshoro were calibrated in the Secondary Standard Dosimetry Laboratory (SSDL) at PINSTECH, using the NPL Secondary Standard Therapy level X-ray exposure meter. The results are presented in this report. (authors)

  6. 42 CFR 423.168 - Accreditation organizations.

    Science.gov (United States)

    2010-10-01

    ..., validation review, onsite observation, or CMS's daily experience with the accreditation organization suggests... whenever it is considering granting an accreditation organization's application for approval. The notice... Federal Register indicating whether it has granted the accreditation organization's request for...

  7. Biological dosimetry - Dose estimation method using biomakers

    International Nuclear Information System (INIS)

    The individual radiation dose estimation is an important step in the radiation risk assessment. In case of radiation incident or radiation accident, sometime, physical dosimetry method can not be used for calculating the individual radiation dose, the other complement method such as biological dosimetry is very necessary. This method is based on the quantitative specific biomarkers induced by ionizing radiation, such as dicentric chromosomes, translocations, micronuclei... in human peripheral blood lymphocytes. The basis of the biological dosimetry method is the close relationship between the biomarkers and absorbed dose or dose rate; the effects of in vitro and in vivo are similar, so it is able to generate the calibration dose-effect curve in vitro for in vivo assessment. Possibilities and perspectives for performing biological dosimetry method in radiation protection area are presented in this report. (author)

  8. Radiation dosimetry and standards at the austrian dosimetry laboratory

    International Nuclear Information System (INIS)

    The Austrian Dosimetry Laboratory, established and operated in cooperation between the Austrian Research Center Seibersdorf and the Federal Office of Metrology and Surveying (Bundesamt and Eich- und Vermessungswesen) maintains the national primary standards for radiation dosimetry. Furthermore its tasks include routine calibration of dosemeters and dosimetric research. The irradiation facilities of the laboratory comprise three X-ray machines covering the voltage range from 5 kV to 420 kV constant potential, a 60Co teletherapy unit, a circular exposure system for routine batch calibration of personnel dosemeters with four gamma ray sources (60Co and 137Cs) and a reference source system with six gamma ray sources (60Co and 137Cs). In addition a set of calibrated beta ray sources are provided (147Pm, 204Tl and 90Sr). The dosimetric equipment consists of three free-air parallelplate ionization chambers serving as primary standards of exposure for the X-ray energy region, graphite cavity chambers with measured volume as primary standards for the gamma radiation of 137Cs and 60Co as well as different secondary standard ionization chambers covering the dose rate range from the natural background level up to the level of modern therapy accelerators. In addition for high energy photon and electron radiation a graphite calorimeter is provided as primary standard of absorbed dose. The principle experimental set-ups for the practical use of the standards are presented and the procedures for the calibration of the different types of dosemeters are described. (Author)

  9. Third conference on radiation protection and dosimetry. Program and abstracts

    Energy Technology Data Exchange (ETDEWEB)

    1991-12-31

    This conference has been designed with the objectives of promoting communication among applied, research, regulatory, and standards personnel involved in radiation protection and providing them with sufficient information to evaluate their programs. To partly fulfill these objectives, a technical program consisting of more than 75 invited and contributed oral presentations encompassing all aspects of radiation protection has been prepared. General topics include external dosimetry, internal dosimetry, instruments, regulations and standards, accreditation and test programs, research advances, and applied program experience. This publication provides a summary of the technical program and a collection of abstracts of the oral presentations.

  10. Third conference on radiation protection and dosimetry. Program and abstracts

    Energy Technology Data Exchange (ETDEWEB)

    None

    1991-01-01

    This conference has been designed with the objectives of promoting communication among applied, research, regulatory, and standards personnel involved in radiation protection and providing them with sufficient information to evaluate their programs. To partly fulfill these objectives, a technical program consisting of more than 75 invited and contributed oral presentations encompassing all aspects of radiation protection has been prepared. General topics include external dosimetry, internal dosimetry, instruments, regulations and standards, accreditation and test programs, research advances, and applied program experience. This publication provides a summary of the technical program and a collection of abstracts of the oral presentations.

  11. Technical basis document for internal dosimetry

    CERN Document Server

    Hickman, D P

    1991-01-01

    This document provides the technical basis for the Chem-Nuclear Geotech (Geotech) internal dosimetry program. Geotech policy describes the intentions of the company in complying with radiation protection standards and the as low as reasonably achievable (ALARA) program. It uses this policy and applicable protection standards to derive acceptable methods and levels of bioassay to assure compliance. The models and computational methods used are described in detail within this document. FR-om these models, dose- conversion factors and derived limits are computed. These computations are then verified using existing documentation and verification information or by demonstration of the calculations used to obtain the dose-conversion factors and derived limits. Recommendations for methods of optimizing the internal dosimetry program to provide effective monitoring and dose assessment for workers are provided in the last section of this document. This document is intended to be used in establishing an accredited dosi...

  12. Survey of international personnel radiation dosimetry programs

    International Nuclear Information System (INIS)

    In September of 1983, a mail survey was conducted to determine the status of external personnel gamma and neutron radiation dosimetry programs at international agencies. A total of 130 agencies participated in this study including military, regulatory, university, hospital, laboratory, and utility facilities. Information concerning basic dosimeter types, calibration sources, calibration phantoms, corrections to dosimeter responses, evaluating agencies, dose equivalent reporting conventions, ranges of typical or expected dose equivalents, and degree of satisfaction with existing systems was obtained for the gamma and neutron personnel monitoring programs at responding agencies. Results of this survey indicate that to provide the best possible occupational radiation monitoring programs and to improve dosimetry accuracy in performance studies, facility dosimetrists, regulatory and standards agencies, and research laboratories must act within their areas of responsibility to become familiar with their radiation monitoring systems, establish common reporting guidelines and performance standards, and provide opportunities for dosimetry testing and evaluation. 14 references, 10 tables

  13. 42 CFR 488.8 - Federal review of accreditation organizations.

    Science.gov (United States)

    2010-10-01

    ... organization proposes to adopt new requirements or change its survey process. An accreditation organization... national accreditation organization. CMS's review and evaluation of a national accreditation organization... criteria— (1) The equivalency of an accreditation organization's accreditation requirements of an entity...

  14. Quality management system of secondary standards dosimetry laboratory in Sri Lanka

    International Nuclear Information System (INIS)

    Full text: Application of Quality Management System (QMS) of Secondary Standard Dosimetry Laboratory (SSDL) of the Atomic Energy Authority (ALA) of Sri Lanka provides path of workflow and information on laboratory operations, management and competence of staff that would assist the laboratory in continual improvement of its processes and meeting accreditation requirements in compliance with IS017025. Thus provision of customers' satisfied accredited dosimetry calibration services is needed for the country. The SSDL currently possesses a reference electrometer (PTW Unidos) with protection level ion- chambers (NE2575, 600cc ion-chamber and PTW - lOLt ion-chamber) and therapy level ion-chambers (NE2571, 0.6cc thimble ion-chamber). Also the laboratory is also having measuring standards (NE2570 electrometer with NE2575, 600cc ion-chamber and NE2571, 0.6cc thimble ion-chamber) . A gamma irradiator which contains two gamma sources (Co-60 and Cs-137) and a X-ray system with six ISO 4037 beam qualities (narrow spectrum of energy range: 33keV - 118keV) are available for protection level X-ray calibrations. Stability of the electrometers with Ion- chambers is performed with Sr-90 check sources, which are specially designed for each type of chambers in order to fix the set-up maintaining the same geometry for every measurement. An average of reading of ten consecutive measurements of which each measurement was made for 300s is taken for stability measurement. Each reading is corrected for ambient temperature and pressure. Acceptance of percentage deviation of stability results with respect to reference reading of respective chamber is ±1% for protection level and ± 0.5% for therapy level. All these equipments, when they are not in used are kept in a dry cabinet in order to control humidity. The SSDL of AEA has become a part of an international network of dosimetry laboratories established by the International Atomic Energy Agency (IAEA) and the World Health Organization (WHO

  15. Dosimetry of total body irradiation

    International Nuclear Information System (INIS)

    In the treatment of disseminated malignancies an improvement in the curability and reduction of complication rates require high precision total body irradiation (TBI) and correct reporting of relevant treatment parameters. Optimal TBI dosimetry is the basis. Radiooncological and radiobiological requirements as well as the special physical situation have to be considered. To review the efforts of medical physicists, highlights from TBI workshops and publications are summarized. Additionally, dosimetric data from 34 European radiooncological centres contributing to the recent ESTRO inquiry on TBI are analysed. The topics are: absorbed dose and dose monitor calibration, determination of absolute and relative doses, dose ratios, attenuation data and heterogeneity corrections; TBI dose calculation methods regarding patient position, beam incidence, body shape and thickness, lung size and density; methods of TBI treatment planning including calculated dose modification and of TBI quality assurance. In conclusion, the following recommendations can be given: TBI dosimetry shall be performed under TBI conditions, close to the real treatment situation. The absorbed dose to water must be determined. The dose monitor should be calibrated against dose measurements at the centre of a water equivalent phantom of TBI equivalent size and typical thickness. Photon fluence profiles have to be measured with small phantoms. Influences on the local dose must be investigated systematically. A reproducible AP/PA TBI technique should be used. The TBI dose shall be specified to mid-abdomen and reported in units of gray. The single and total dose and the dose rate to the lungs, the number of fractions and the treatment time schedule must be stated. In vivo dosimetry is required if non-reliable TBI techniques are used. An international TBI dosimetry intercomparison could assist these efforts to improve the treatment of acute leukaemia. (author). 89 refs, 3 figs, 13 tabs

  16. Feasibility of using neural networks to unfold the response of multi-element TLD for mixed field dosimetry

    International Nuclear Information System (INIS)

    Significant advances have been made in recent years to improve calibration methodology and dose calculation algorithm in the fields of TL dosimetry. This process was accelerated in the past decade particularly in the Republic of Korea by the need to meet mandatory national accreditation requirements. The objective of this study is to develop a new algorithm to replace the simplistic decision tree algorithms by the more sophisticated neural networks in hopes of achieving a higher degree of accuracy and precision in personnel dosimetry system. The original hypothesis of this work is that the spectral information of an X and γ-ray fields may be obtained by the analysis of the response of a multi-element system. In this study, a feed forward neural network using the error back-propagation method with Bayesian optimization was designed for the response unfolding procedure. The response functions of the single element to photons were calculated by application of a computational Monte-Carlo model for an energy range from 10 keV to 2 MeV with different spectral proportions. The training of the artificial neural network was based on the computation of responses of a four-element system for the back-propagation method. The validation of the proposed algorithm was investigated by unfolding the 10 computed responses for arbitrary mixed gamma fields and the spectra resulting from the unfolding procedure agree well with the original spectra. (author)

  17. Calibration of sources for alpha spectroscopy systems

    International Nuclear Information System (INIS)

    This paper describes the calibration methodology for measuring the total alpha activity of plane and thin sources with the Alpha Spectrometer for Silicon Detector in the Nuclear Measures and Dosimetry laboratory at IEAv/CTA. (author)

  18. Distinctions among Accreditation Agencies for Business Programs

    Science.gov (United States)

    Corcoran, Charles P.

    2007-01-01

    Over the past twenty years, business accreditation has become a growth industry. In 1988, some eleven percent of business programs were accredited by an accrediting body devoted solely to business program accreditation. Today, over forty-two percent boast of such external validation of their programs. Although the three principal accrediting…

  19. 7 CFR 983.1 - Accredited laboratory.

    Science.gov (United States)

    2010-01-01

    ... 7 Agriculture 8 2010-01-01 2010-01-01 false Accredited laboratory. 983.1 Section 983.1 Agriculture..., ARIZONA, AND NEW MEXICO Definitions § 983.1 Accredited laboratory. An accredited laboratory is a laboratory that has been approved or accredited by the U.S. Department of Agriculture....

  20. CAECC Software Testing Laboratory Accredited by CNAL

    Institute of Scientific and Technical Information of China (English)

    2005-01-01

    Software Testing Laboratory of China Aerospace Engineering and Consultation Center (CAECC) is accredited by China National Accreditation Board for Laboratories (CNAL) as the first such laboratory in domestic space industry. Since CNAL is a member of International Laboratory Accreditation Cooperation (ILAC),software testing reports certificated to CAECC are recognized by 45 laboratory accreditation organizations in AsiaPacific region, Europe and America.

  1. Dosimetry for radiation processing

    DEFF Research Database (Denmark)

    Miller, Arne

    1986-01-01

    During the past few years significant advances have taken place in the different areas of dosimetry for radiation processing, mainly stimulated by the increased interest in radiation for food preservation, plastic processing and sterilization of medical products. Reference services both...... and sterilization dosimetry, optichromic dosimeters in the shape of small tubes for food processing, and ESR spectroscopy of alanine for reference dosimetry. In this paper the special features of radiation processing dosimetry are discussed, several commonly used dosimeters are reviewed, and factors leading...

  2. SPRT Calibration Uncertainties and Internal Quality Control at a Commercial SPRT Calibration Facility

    Science.gov (United States)

    Wiandt, T. J.

    2008-06-01

    The Hart Scientific Division of the Fluke Corporation operates two accredited standard platinum resistance thermometer (SPRT) calibration facilities, one at the Hart Scientific factory in Utah, USA, and the other at a service facility in Norwich, UK. The US facility is accredited through National Voluntary Laboratory Accreditation Program (NVLAP), and the UK facility is accredited through UKAS. Both provide SPRT calibrations using similar equipment and procedures, and at similar levels of uncertainty. These uncertainties are among the lowest available commercially. To achieve and maintain low uncertainties, it is required that the calibration procedures be thorough and optimized. However, to minimize customer downtime, it is also important that the instruments be calibrated in a timely manner and returned to the customer. Consequently, subjecting the instrument to repeated calibrations or extensive repeated measurements is not a viable approach. Additionally, these laboratories provide SPRT calibration services involving a wide variety of SPRT designs. These designs behave differently, yet predictably, when subjected to calibration measurements. To this end, an evaluation strategy involving both statistical process control and internal consistency measures is utilized to provide confidence in both the instrument calibration and the calibration process. This article describes the calibration facilities, procedure, uncertainty analysis, and internal quality assurance measures employed in the calibration of SPRTs. Data will be reviewed and generalities will be presented. Finally, challenges and considerations for future improvements will be discussed.

  3. Neutron dosimetry - A review

    International Nuclear Information System (INIS)

    This review summarizes information on the following subjects: (1) physical processes of importance in neutron dosimetry; (2) biological effects of neutrons; (3) neutron sources; and (4) instruments and methods used in neutron dosimetry. Also, possible improvements in dosimetry instrumentation are outlined and discussed. (author)

  4. [Accreditation of processes in hepatology].

    Science.gov (United States)

    Clemente, Gerardo; Pérez-Lázaro, Juan José; Tejedor, Martín; Planas, Ramón; De la Mata, Manuel; Córdoba, Juan; Jara, Paloma; Herrero, José Ignacio; Prieto, Martín; Suáreza, Gonzalo; Arroyo, Vicente

    2008-01-01

    The Spanish Association for the Study of the Liver decided in 2006 to develop a project to assess the quality of the professionals, processes and medical units dealing with the management of patients with liver diseases in Spain. The current article reports the criteria proposed to assess the quality and the accreditation of the processes in hepatology. The processes considered include most patients with liver diseases and the accreditation system designed is highly specific. This document, together with a previous one published in gastroenterología y hepatología concerning the accreditation of the professionals and a third document dealing with the accreditation of liver units that will be published soon, form the basis of the quality assessment of hepatology in our country.

  5. Secondary standard dosimetry laboratory at INFLPR

    Energy Technology Data Exchange (ETDEWEB)

    Scarlat, F.; Minea, R.; Scarisoreanu, A.; Badita, E.; Sima, E.; Dumitrascu, M.; Stancu, E.; Vancea, C., E-mail: scarlat.f@gmail.com [National Institute for Laser, Plasma and Radiation Physics - INFLPR, Bucharest (Romania)

    2011-07-01

    National Institute for Laser, Plasma and Radiation Physics (INFLPR) has constructed a High Energy Secondary Standard Dosimetry Laboratory SSDL-STARDOOR - for performing dosimetric calibrations according to ISO IEC SR/EN 17025:2005 standards. This is outfitted with UNIDOS Secondary Standard Dosimeter from PTW (Freiburg Physikalisch-Technische Werksttaten) calibrated at the PTB-Braunschweig (German Federal Institute of Physics and Metrology). A radiation beam of the quality of Q used by our laboratory as calibration source are provided by INFLPR 7 MeV electron beam linear accelerator mounted in our facility. (author)

  6. Perspective from a commercial supplier of dosimetry services

    International Nuclear Information System (INIS)

    The traditional radiation related industries in the United States have matured. The growth rates in the numbers of radiation workers have moderated and ALARA programs have favorably reduced many exposures. Dosimetry testing and accreditation by the National Bureau of Standards have identified those services possessing satisfactory dosimetry systems and technical competence. These developments have influenced the business perspectives. Combined with the overall renewed emphasis on competition and productivity in American business, many dosimetry services have become more aggressive in seeking new markets; residential radon measurements being most obvious. The potential size of these markets is making investments in technical research more attractive. In the past, most research funding was provided by the government. The renewed research interest by the private sector could stimulate the entry of new professionals into radiation measurement research. Research results have the potential for improving traditional services and expanding the applicability of certain measurement methods

  7. Dosimetry on the radiological risks prevention in radiotherapy

    International Nuclear Information System (INIS)

    Dosimetry in its various forms plays a determining role on the radiological risks prevention in radiotherapy. To prove this in this paper is shown an analysis based on the risk matrix method, how the dosimetry can influence in each stages of a radiotherapy service; installation and acceptance, operation, maintenance and calibration. For each one of these stages the role that can play is analyzed as either the initiating event of a radiological accident or limiting barrier of these events of the dosimetric processes used for the individual dosimetry, the area monitoring, fixed or portable, for radiation beam dosimetry and of the patients for a radiotherapy service with cobalt-therapy equipment. The result of the study shows that the application of a prospective approach in the role evaluation of dosimetry in the prevention and mitigation of the consequences of a radiological accident in radiotherapy is crucial and should be subject to permanent evaluation at each development stage of these services. (author)

  8. Performance testing of UK personal dosimetry laboratories

    CERN Document Server

    Marshall, T O

    1985-01-01

    The proposed Ionising Radiations Regulations will require all UK personal dosimetry laboratories that monitor classified personnel to be approved for personal dosimetry by the Health and Safety Executive. It is suggested that these approvals should be based on general and supplementary criteria published by the British Calibration Service (BCS) for laboratory approval for the provision of personal dosimetry services. These criteria specify certain qualitative requirements and also indicate the need for regular tests of performance to be carried out to ensure constancy of dosimetric standards. This report concerns the latter. The status of the BCS criteria is discussed and the need for additional documents to cover new techniques and some modifications to existing documents is indicated. A means is described by which the technical performance of laboratories, concerned with personal monitoring for external radiations, can be assessed, both initially and ongoing. The costs to establish the scheme and operate it...

  9. Instrumentation for the individual dosimetry of workers

    CERN Document Server

    Thévenin, J C

    2003-01-01

    The control of the radiation dose exposure of workers and personnel exposed to ionizing radiations (nuclear industry, nuclear medicine, army, university laboratories etc..) is ensured by individual dosemeters. This dosimetry is mandatory for all workers susceptible to be exposed to more than 30% of the regulatory dose limit. dosemeters are worn on the chest and in some particular cases, on the finger (dosemeter rings) or on the wrist. Passive dosemeters allow to measure the dose a posteriori, while electronic dosemeters allow a direct reading and recording of the dose. This article presents successively: 1 - the general principles of individual dosimetry: situations of exposure, radiation detection, operational data, standardization, calibration and quality assurance, measurement uncertainties; 2 - goals and regulatory framework of individual dosimetry: regulation and recommendations, optimization, respect of dose limits, accidental situations; 3 - passive dosemeters: film, thermoluminescent, radio-photolumin...

  10. Statistical analysis of personal dosimetry of exposed workers; Analisis estadistico de la dosimetria personal de trabajadores expuestos

    Energy Technology Data Exchange (ETDEWEB)

    Sanchez Munoz, F. J.; Alejo Luque, L.; Mas Munoz, I.; Serrada Hierro, A.

    2013-07-01

    The dosimetry centers accredited by the Nuclear Safety Council (CSN) normally report overcoming legal limits, or some fraction thereof, but do not provide comparative dosimetric criteria indicating if assigned to a given dose is large TPE or small relative to that of their peers. In order to help to resolve the difficulties mentioned ds, it has developed an application that statistically processes the dosimetric data provided by the National Dosimetry Center. (Author)

  11. HSE performance tests for dosimetry services

    International Nuclear Information System (INIS)

    In the United Kingdom a dosimetry service that measures and assesses whole-body or part-body doses arising from external radiation must successfully complete a performance test. Results of the performance tests for routine whole-body, routine extremity/skin and special accident dosimetry, carried out over the past six years by the AEA Technology Calibration Service at Winfrith, and DRaStaC, the AWE Calibration Service at Aldermaston, are presented. The test involves irradiating groups of dosemeters to known doses of gamma radiation and determining the bias and relative standard deviations for each dose group. The results are compared with the pass criteria specified by the UK Health and Safety Executive. For routine whole-body dosimetry, both the film badge and thermoluminescent dosemeter (TLD) perform adequately for irradiations between 0.6 and 30 mSv. For higher doses up to 250 mSv, where the slow emulsion of the film is used, the film badge shows poorer performance with a tendency to overestimate the dose. For routine extremity/skin dosimetry there is a wider spread of relative standard deviation results than is seen for routine whole-body dosimetry. This is to be expected since the results will include dosemeters that are based on 'disposable' TLDs and ones based on lithium fluoride powder in sachets. For special accident dosimetry the dosemeters are tested between 0.26 and 6 Gy. For the highest dose group the film badge invariably underestimates the true dose, whereas the TLD has a tendency to overestimate it. (author)

  12. General guidance for laboratories providing personal dosimetry services

    International Nuclear Information System (INIS)

    This guidance is recommended to all dosimetry services in the interests of good radiation protection practice. For dosimetry services who seek approval under the Ionising Radiation Regulations 1985, the Health and Safety Executive (HSE) would invoke compliance with this guidance, in broad terms, as well as other published guidance. The recommendations include sections concerning laboratory organizations and staff, documentation of procedures, laboratory accommodation and services, equipment, calibration and traceability, housekeeping, dosimetric capability, laboratory records. (Author)

  13. Radiation Dosimetry for Quality Control of Food Preservation and Disinfestation

    DEFF Research Database (Denmark)

    McLaughlin, W.L.; Miller, Arne; Uribe, R.M.

    1983-01-01

    In the use of x and gamma rays and scanned electron beams to extend the shelf life of food by delay of sprouting and ripening, killing of microbes, and control of insect population, quality assurance is provided by standardized radiation dosimetry. By strategic placement of calibrated dosimeters...... speed) to meet changes that occur in product and source parameters (e.g. bulk density and radiation spectrum). Routine dosimetry methods and certain corrections of dosimetry data may be selected for the radiations used in typical food processes....

  14. Radiation Dosimetry for Quality Control of Food Preservation and Disinfestation

    DEFF Research Database (Denmark)

    McLaughlin, W.L.; Miller, Arne; Uribe, R.M.

    1983-01-01

    In the use of x and gamma rays and scanned electron beams to extend the shelf life of food by delay of sprouting and ripening, killing of microbes, and control of insect population, quality assurance is provided by standardized radiation dosimetry. By strategic placement of calibrated dosimeters...... parameters (e.g. conveyor speed) to meet changes that occur in product and source parameters (e.g. bulk density and radiation spectrum). Routine dosimetry methods and certain corrections of dosimetry data may be selected for the radiations used in typical food processes....

  15. Dosimetry in radiation processing- Indian scenario

    International Nuclear Information System (INIS)

    Radiation processing is a method for producing chemical, physical, and microbiological changes in substances by exposing to ionizing radiation. Availability of high intensity cobalt-60 gamma ray sources and high power electron beam accelerators has led to a continuous growth of radiation processing industry in India. Commercial viability and safe operation of these radiation-processing plants depends on accurate dosimetry. Depending on the purpose to be achieved, a widespread dose range, from few grays to few hundred kilo grays, is encountered in radiation processing technology and this necessitates the use of different dosimetry systems. In the present paper, current status of radiation processing facilities in India has been reviewed. Various indigenously developed dosimetry systems such as Alanine/glutamine (Spectrophotometric readout), FBX and ceric-cerous (potentiometry) are being used for quality assurance and routine plant dosimetry. Fricke dosimeter is used as a reference standard for calibrating other dosimetry systems. Glutamine (Spectrophotometric read out) dosimeter, used as transfer standard for Q.A. has traceability to NPL, UK and has shown an agreement within ±2% during dose intercomparisons carried out with various international standards laboratories. Performance of these dosimeters was found to be better than ±10% during dose measurements in radiation sterilization and food irradiation plants. (author)

  16. Standard Guide for Performance Characterization of Dosimeters and Dosimetry Systems for Use in Radiation Processing

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2009-01-01

    1.1 This guide provides guidance on determining the performance characteristics of dosimeters and dosimetry systems used in radiation processing. 1.2 This guide describes the influence quantities that might affect the performance of dosimeters and dosimetry systems and that should be considered during dosimeter/dosimetry system characterization. 1.3 Users of this guide are directed to existing standards and literature for procedures to determine the effects from individual influence quantities and from combinations of more than one influence quantity. 1.4 Guidance is provided regarding the roles of the manufacturers, suppliers, and users in the characterization of dosimeters and dosimetry systems. 1.5 This guide does not address how the dosimeter/dosimetry system characterization information is to be used in radiation processing applications or in the calibration of dosimetry systems. Note 1—For guidance on the use of dosimeter/dosimetry system characterization information for the selection and use o...

  17. 7 CFR 205.506 - Granting accreditation.

    Science.gov (United States)

    2010-01-01

    ... Agriculture Regulations of the Department of Agriculture (Continued) AGRICULTURAL MARKETING SERVICE (Standards, Inspections, Marketing Practices), DEPARTMENT OF AGRICULTURE (CONTINUED) ORGANIC FOODS PRODUCTION ACT PROVISIONS NATIONAL ORGANIC PROGRAM Accreditation of Certifying Agents § 205.506 Granting accreditation....

  18. Accreditation of Engineering Programs In The USA

    OpenAIRE

    Jones, E. C.; Reyes-Guerra, David R.

    1989-01-01

    Accreditation is a way of assessing the quality of education. In the United States accreditation of engineering programs is carried out by volunteers, engineering educators and practitioners who evaluate programs against criteria developed by the profession. Universities voluntarily submit their engineering programs for professional accreditation. The process is supported by various professional engineering societies, the universities that request accreditation, the volunteers who carry out t...

  19. The Accreditation of Laboratories Proficiency and Safety

    Institute of Scientific and Technical Information of China (English)

    2005-01-01

    @@ Recently, China National Accreditation Board for Laboratories (CNAL) has released CNAL/AC23:2004 Medical Laboratories: Accreditation Criteria For Quality and Proficiency, and meanwhile GB 19489 Laboratories: General Requirements For Biosafety and ISO 15190 Medical Laboratories-Requirements For Safetywill be adopted by CNAL as the accreditation criteria for laboratories safety.

  20. Dosimetry intercomparisons between fast neutron radiotherapy facilities

    International Nuclear Information System (INIS)

    Neutron dosimetry intercomparisons have been made between M.D. Anderson Hospital and Tumor Institute, Naval Research Laboratory, University of Washington Hospital, and Hammersmith Hospital. The parameters that are measured during these visits are: tissue kerma in air, tissue dose at depth of dose maximum, depth dose, beam profiles, neutron/gamma ratios and photon calibrations of ionization chambers. A preliminary report of these intercomparisons will be given including a comparison of the calculation and statement of tumor doses for each institution

  1. The Accreditation-Eligibility Link.

    Science.gov (United States)

    Levin, Nora Jean

    1981-01-01

    Public policy rests on the unreliable assumption that postsecondary education institutions and programs approved by nationally recognized private accrediting agencies are bona fide providers of educational services, worthy of students' time, effort, and money and of federal funds. Rather, federal fund eligibility should focus on measures of…

  2. Accreditation of Industrial Engineering Programs.

    Science.gov (United States)

    Brooks, George H.

    The guidelines used in the accreditation of industrial engineering programs are discussed. Changes that have taken place in engineering curriculum are described, along with the philosophy of educators in formulating industrial engineering program requirements in the areas of faculty, facilities, curriculum, administration, and scholastic work.…

  3. Specialized Accreditation: College Library Responses.

    Science.gov (United States)

    Frazer, Stuart

    1994-01-01

    Offers a detailed overview of accreditation standards, criteria, and procedures used by one specialized accreditor; and suggests ways to be prepared for site visits and reports. A directory of 73 specialized accreditors and a brief bibliography are appended. (18 references) (Author/SLW)

  4. Dosimetry service removal

    CERN Multimedia

    Safety Commission

    2010-01-01

    Dear personal dosimeter user, Please note that the Dosimetry service has moved in building 55, the service is now located in the main floor: 55-R-004. Main floor instead of second floor. On your right hand when accessing in the building. Thank you Dosimetry Service

  5. Dosimetry in process control

    International Nuclear Information System (INIS)

    Measurement of absorbed dose and dose distribution in irradiated medical products relies on the use of quality dosimetry systems, trained personnel and a thorough understanding of the energy deposition process. The interrelationship of these factors will be discussed with emphasis on the current and future practices of process control dosimetry. (author)

  6. Accreditation to supervise research

    International Nuclear Information System (INIS)

    In this document the author reviews his works between 1995 and 2010. First, the development of a silicon pixel detector is detailed, the purpose of this detector was to improve the forward proton spectrometer of the H1 experiment at DESY. The works made to develop the reading circuits of the pixel detector are presented, particularly the design of the test bench for the testing of these circuits and the simulation of their behaviour in realistic environment. The second part describes the design of the front electronic for the data acquisition of the calorimeter detector of ATLAS (TileCal) and its testing system (MobiDICK). The software for the control system of the laser calibration of TileCal is detailed. The last part gives an account of the author's activities in the field of science popularization through the 'Cosmophone' and knowledge dissemination. The Cosmophone is a particle detector that turns the passage of particles into sounds in order to make the general public more aware of the presence of particles

  7. IAEA Support to National TLD Audit Networks for Radiotherapy Dosimetry

    International Nuclear Information System (INIS)

    For several years, the IAEA has supported the development of methodology and establishment of national quality audit networks for radiotherapy dosimetry. The main objective was to extend the availability of radiotherapy dosimetry audits to as many radiotherapy centres as possible throughout the world. Since 1995, a series of three coordinated research projects (CRPs) has been conducted by the IAEA to assist its Member States to develop such national audit programmes. The first CRP focused on the basic beam calibration audits. The basic programme was extended to audits in non-reference conditions through a second CRP. The third CRP initiated in 2009 is expanding the dosimetry audit tools for more complex techniques used for treatment of cancer patients. The national audit networks participating in these CRPs have incorporated in their programmes procedures for auditing hospital dosimetry for these techniques. (author)

  8. INPO accreditation - product definition for utility training

    International Nuclear Information System (INIS)

    As a part of its responsibility to promote training excellence, the Institute of Nuclear Power Operations (INPO) initiated an accreditation program in 1982 on behalf of its member utilities. The purpose of the accreditation program is to assist INPO members in developing and maintaining training programs that produce well-qualified, competent personnel to operate the nation's nuclear power plants. Accreditation formally recognizes nuclear utility training as meeting the INPO accreditation objectives and criteria for initial and continuing training programs for operations, maintenance, and technical personnel. The ultimate objective to be achieved by accreditation is to maintain high-quality training and enhance the professionalism of the personnel who operate the nuclear power plants. While initial accreditation represents recognition that quality training programs have been put in place at the nuclear power plants, the renewal of accreditation represents recognition that quality training programs have been put in place at the nuclear power plants, the renewal of accreditation will keep the training programs vital, effective, and up to high standards. The nuclear power industry has accepted the responsibility of striving for excellence in training its personnel to safely operate the power plants. The full use of accreditation and the accrediting process is an important means to fulfilling this responsibility

  9. The United Kingdom's radiotherapy dosimetry audit network

    International Nuclear Information System (INIS)

    The first comprehensive national dosimetry intercomparison in the United Kingdom involving all UK radiotherapy centres was carried out in the late 1980s. Out of this a regular radiotherapy dosimetry audit network evolved in the early 1990s. The network is co-ordinated by the Institute of Physics and Engineering in Medicine and comprises eight co-operative regional groups. Audits are based on site visits using ionization chambers and epoxy resin water substitute phantoms. The basic audit methodology and phantom design follows that of the original national intercomparison exercise. However, most of the groups have evolved more complex methods, to extend the audit scope to include other parameters, other parts of the radiotherapy process and other treatment modalities. A number of the groups have developed phantoms to simulate various clinical treatment situations, enabling the sharing of phantoms and expertise between groups, but retaining a common base. Besides megavoltage external beam photon dosimetry, a number of the groups have also included the audit of kilovoltage X ray beams, electron beams and brachytherapy dosimetry. The National Physical Laboratory is involved in the network and carries out basic beam calibration audits to link the groups. The network is described and the methods and results are illustrated using the Scottish+ group as an example. (author)

  10. In vivo dosimetry with silicon diodes in total body irradiation

    International Nuclear Information System (INIS)

    The aim of this work is the characterization and application of silicon diode detectors for in vivo dosimetry in total body irradiation (TBI) treatments. It was evaluated the diode response with temperature, dose rate, gantry angulations and field size. A maximum response variation of 2.2% was obtained for temperature dependence. The response variation for dose rate and angular was within 1.2%. For field size dependence, the detector response increased with field until reach a saturation region, where no more primary radiation beam contributes for dose. The calibration was performed in a TBI setup. Different lateral thicknesses from one patient were simulated and then the calibration factors were determined by means of maximum depth dose readings. Subsequent to calibration, in vivo dosimetry measurements were performed. The response difference between diode readings and the prescribed dose for all treatments was below 4%. This difference is in agreement as recommended by the International Commission on Radiation Units and Measurements (ICRU), which is ±5%. The present work to test the applicability of a silicon diode dosimetry system for performing in vivo dose measurements in TBI techniques presented good results. These measurements demonstrated the value of diode dosimetry as a treatment verification method and its applicability as a part of a quality assurance program in TBI treatments. - Highlights: ► Characterization of a silicon diode dosimetry system. ► Application of the diodes for in vivo dosimetry in total body irradiation treatments. ► Implementation of in vivo dosimetry as a part of a quality assurance program in radiotherapy

  11. Intercomparison exercise within a distributed-dosimetry network

    International Nuclear Information System (INIS)

    The results of an intercomparison exercise within the US Navy dosimetric network (USN-DN) are presented and discussed. The USN-DN uses a commercially available LiF:Mg,Cu,P thermoluminescent dosemeter (TLD) model Harshaw 8840/8841 and TLD reader model Harshaw 8800 manufactured by Thermo Fisher Scientific. The USN-DN consists of a single calibration facility and 16 satellite dosimetry reading facilities throughout the world with ∼40 model 8800 TLD readers and in excess of 350 000 TLD cards in circulation. The Naval Dosimetry Center (NDC) is the primary calibration site responsible for the distribution and calibration of all TLD cards and their associated holders. In turn, each satellite facility is assigned a subpopulation of cards, which are utilised for servicing their local customers. Consistency of the NDC calibration of 150 dosemeters (calibrated at NDC) and 27 locally calibrated remote readers was evaluated in the framework of this intercomparison. Accuracy of TLDs' calibration, performed at the NDC, was found to be <3 % throughout the entire network. Accuracy of the readers' calibration, performed with the NDC issued calibration dosemeters at remote sites, was found to be better than 4 % for most readers. The worst performance was found for reader Channel 3, which is calibrated using the thinnest chip of the Harshaw 8840/8841 dosemeter. The loss of sensitivity of this chip may be caused by time-temperature profile that has been designed for all four chips without consideration of chip thickness. (authors)

  12. [ISO 15189 medical laboratory accreditation].

    Science.gov (United States)

    Aoyagi, Tsutomu

    2004-10-01

    This International Standard, based upon ISO/IEC 17025 and ISO 9001, provides requirements for competence and quality that are particular to medical laboratories. While this International Standard is intended for use throughout the currently recognized disciplines of medical laboratory services, those working in other services and disciplines will also find it useful and appropriate. In addition, bodies engaged in the recognition of the competence of medical laboratories will be able to use this International Standard as the basis for their activities. The Japan Accreditation Board for Conformity Assessment (AB) and the Japanese Committee for Clinical Laboratory Standards (CCLS) are jointly developing the program of accreditation of medical laboratories. ISO 15189 requirements consist of two parts, one is management requirements and the other is technical requirements. The former includes the requirements of all parts of ISO 9001, moreover it includes the requirement of conformity assessment body, for example, impartiality and independence from any other party. The latter includes the requirements of laboratory competence (e.g. personnel, facility, instrument, and examination methods), moreover it requires that laboratories shall participate proficiency testing(s) and laboratories' examination results shall have traceability of measurements and implement uncertainty of measurement. Implementation of ISO 15189 will result in a significant improvement in medical laboratories management system and their technical competence. The accreditation of medical laboratory will improve medical laboratory service and be useful for patients. PMID:15624503

  13. [ISO 15189 medical laboratory accreditation].

    Science.gov (United States)

    Aoyagi, Tsutomu

    2004-10-01

    This International Standard, based upon ISO/IEC 17025 and ISO 9001, provides requirements for competence and quality that are particular to medical laboratories. While this International Standard is intended for use throughout the currently recognized disciplines of medical laboratory services, those working in other services and disciplines will also find it useful and appropriate. In addition, bodies engaged in the recognition of the competence of medical laboratories will be able to use this International Standard as the basis for their activities. The Japan Accreditation Board for Conformity Assessment (AB) and the Japanese Committee for Clinical Laboratory Standards (CCLS) are jointly developing the program of accreditation of medical laboratories. ISO 15189 requirements consist of two parts, one is management requirements and the other is technical requirements. The former includes the requirements of all parts of ISO 9001, moreover it includes the requirement of conformity assessment body, for example, impartiality and independence from any other party. The latter includes the requirements of laboratory competence (e.g. personnel, facility, instrument, and examination methods), moreover it requires that laboratories shall participate proficiency testing(s) and laboratories' examination results shall have traceability of measurements and implement uncertainty of measurement. Implementation of ISO 15189 will result in a significant improvement in medical laboratories management system and their technical competence. The accreditation of medical laboratory will improve medical laboratory service and be useful for patients.

  14. Dosimetry practices at the Radiation Technology Centre (Ghana)

    International Nuclear Information System (INIS)

    Dosimetry practices undertaken to support research and pilot scale gamma irradiation activities at the Radiation Technology Centre of the Ghana Atomic Energy Commission are presented. The Fricke dosemeter was used for calibrating the gamma field of the gammacell-220. The Fricke system and the gammacell-220 were then used to calibrate the ethanol chlorobenzene (ECB) dosemeter. The Fricke and ECB dosemeter systems have become routine dosemeters at the centre. Dosimetry work has covered a wide range of research specimens and pilot scale products to establish the relevant irradiation protocol and parameters for routine treatment. These include yams, pineapple explants, blood for feeding tsetseflies, cocoa bud wood and cassava sticks. Pilot scale dosimetry studies on maize, medical devices like intravenous infusion sets and surgical gauze have also been completed. The results and observations made on some of these products are reported. (author). 4 refs., 5 figs

  15. Current work on dosimetry standards in Japan

    International Nuclear Information System (INIS)

    Basic concepts on standardization of radiation dosimetry are reviewed. The present situation regarding primary standards in the Electrotechnical Laboratory, the primary standard dosimetry laboratory in Japan, is presented, considering the following; (i) Established dosimetric standards of exposure for soft and medium-energy X-rays and gamma rays. This section includes methods of their standardization, and discusses accuracies of instruments operating as environmental-level, protection-level, inspection-level, therapy-level, and processing-level measuring systems. The results of international comparisons between ETL and other, foreign primary standard dosimetry laboratories are presented; (ii) Other established radiation standards related to derivation of radiation absorbed dose. These primary standards include those for the neutron emission rates, thermal and fast neutron flux densities, energy fluences for high-energy photons and electrons, and activities of several kinds of radioactive material. The accuracies and results of international comparisons relating to them are also presented; (iii) Research being carried out at ETL. The current status of the dissemination of radiation standards is presented considering in particular: (i) The calibration services available at ETL, the categories of these services, energy and dose rate ranges, methods, accuracies, etc.; (ii) The calibration services available in certain other organizations considered as SSDLs in Japan, the categories of such work, methods, accuracies etc.; (iii) Present endeavours towards establishing a systematic and effective dissemination system (a so-called Traceability System) in Japan

  16. Factors influencing EPR dosimetry in fingernails

    International Nuclear Information System (INIS)

    The technique based on the detection of ionizing radiation induced radicals by EPR in tooth enamel is an established method for the dosimetry of exposed persons in radiological emergencies. Dosimetry based on EPR spectral analysis of fingernail clippings, currently under development, has the practical advantage of the easier sample collection. A limiting factor is that overlapping the radiation induced signal (RIS), fingernails have shown the presence of two mechanically induced signals, called MIS1 and MIS2, due to elastic and plastic deformation respectively, at the time of fingernails cutting. With a water treatment, MIS1 is eliminated while MIS2 is considerably reduced. The calibration curves needed for radiation accident dosimetry should have 'universal' characteristics, ie. Represent the variability that can be found in different individuals. Early studies were directed to the analysis of factors affecting the development of such universal calibration curves. The peak to peak amplitude of the signal before and after the water treatment as well as the effect of size and number of clippings were studied. Furthermore, the interpersonal and intrapersonal variability were analyzed. Taking into account these previous studies, the optimal conditions for measurement were determined and EPR spectra of samples irradiated at different doses were used for the developing of dose-response curves. This paper presents the analysis of the results.(authors)

  17. Early experiences of accredited clinical informatics fellowships.

    Science.gov (United States)

    Longhurst, Christopher A; Pageler, Natalie M; Palma, Jonathan P; Finnell, John T; Levy, Bruce P; Yackel, Thomas R; Mohan, Vishnu; Hersh, William R

    2016-07-01

    Since the launch of the clinical informatics subspecialty for physicians in 2013, over 1100 physicians have used the practice and education pathways to become board-certified in clinical informatics. Starting in 2018, only physicians who have completed a 2-year clinical informatics fellowship program accredited by the Accreditation Council on Graduate Medical Education will be eligible to take the board exam. The purpose of this viewpoint piece is to describe the collective experience of the first four programs accredited by the Accreditation Council on Graduate Medical Education and to share lessons learned in developing new fellowship programs in this novel medical subspecialty. PMID:27206458

  18. Engineering Accreditation in China: The Progress and Development of China's Engineering Accreditation

    Science.gov (United States)

    Jiaju, Bi

    2009-01-01

    Among engineering degree programs at the bachelor's level in China, civil engineering was the first one accredited in accordance with a professional programmatic accreditation system comparable to that of international practice. Launched in 1994, the accreditation of civil engineering aimed high and toward international standards and featured the…

  19. Hanford External Dosimetry Technical Basis Manual PNL-MA-842

    Energy Technology Data Exchange (ETDEWEB)

    Rathbone, Bruce A.

    2010-04-01

    The Hanford External Dosimetry Technical Basis Manual PNL-MA-842 documents the design and implementation of the external dosimetry system used at the U.S. Department of Energy (DOE) Hanford site. The manual describes the dosimeter design, processing protocols, dose calculation methodology, radiation fields encountered, dosimeter response characteristics, limitations of dosimeter design under field conditions, and makes recommendations for effective use of the dosimeters in the field. The manual describes the technical basis for the dosimetry system in a manner intended to help ensure defensibility of the dose of record at Hanford and to demonstrate compliance with requirements of 10 CFR 835, the DOE Laboratory Accreditation Program, the DOE Richland Operations Office, DOE Office of River Protection, DOE Pacific Northwest Office of Science, and Hanford’s DOE contractors. The dosimetry system is operated by the Pacific Northwest National Laboratory (PNNL) Hanford External Dosimetry Program which provides dosimetry services to PNNL and all Hanford contractors. The primary users of this manual are DOE and DOE contractors at Hanford using the dosimetry services of PNNL. Development and maintenance of this manual is funded directly by DOE and DOE contractors. Its contents have been reviewed and approved by DOE and DOE contractors at Hanford through the Hanford Personnel Dosimetry Advisory Committee which is chartered and chaired by DOE-RL and serves as means of coordinating dosimetry practices across contractors at Hanford. This manual was established in 1996. Since its inception, it has been revised many times and maintained by PNNL as a controlled document with controlled distribution. The first revision to be released through PNNL’s Electronic Records & Information Capture Architecture database was designated Revision 0. Revision numbers that are whole numbers reflect major revisions typically involving significant changes to all chapters in the document. Revision

  20. Hanford External Dosimetry Technical Basis Manual PNL-MA-842

    Energy Technology Data Exchange (ETDEWEB)

    Rathbone, Bruce A.

    2011-04-04

    The Hanford External Dosimetry Technical Basis Manual PNL-MA-842 documents the design and implementation of the external dosimetry system used at the U.S. Department of Energy (DOE) Hanford site. The manual describes the dosimeter design, processing protocols, dose calculation methodology, radiation fields encountered, dosimeter response characteristics, limitations of dosimeter design under field conditions, and makes recommendations for effective use of the dosimeters in the field. The manual describes the technical basis for the dosimetry system in a manner intended to help ensure defensibility of the dose of record at Hanford and to demonstrate compliance with requirements of 10 CFR 835, the DOE Laboratory Accreditation Program, the DOE Richland Operations Office, DOE Office of River Protection, DOE Pacific Northwest Office of Science, and Hanford’s DOE contractors. The dosimetry system is operated by the Pacific Northwest National Laboratory (PNNL) Hanford External Dosimetry Program which provides dosimetry services to PNNL and all Hanford contractors. The primary users of this manual are DOE and DOE contractors at Hanford using the dosimetry services of PNNL. Development and maintenance of this manual is funded directly by DOE and DOE contractors. Its contents have been reviewed and approved by DOE and DOE contractors at Hanford through the Hanford Personnel Dosimetry Advisory Committee which is chartered and chaired by DOE-RL and serves as means of coordinating dosimetry practices across contractors at Hanford. This manual was established in 1996. Since its inception, it has been revised many times and maintained by PNNL as a controlled document with controlled distribution. The first revision to be released through PNNL’s Electronic Records & Information Capture Architecture database was designated Revision 0. Revision numbers that are whole numbers reflect major revisions typically involving significant changes to all chapters in the document. Revision

  1. IAEA/WHO TLD postal dose audit service and high precision measurements for radiotherapy level dosimetry

    International Nuclear Information System (INIS)

    Since 1969 the International Atomic Energy Agency, together with the World Health Organization, has performed postal TLD audits to verify calibration of radiotherapy beams in developing countries. The TLD programme also monitors activities of Secondary Standard Dosimetry Laboratories (SSDLs). The programme has checked approximately 4000 clinical beams in over 1100 hospitals, and in many instances significant errors have been detected in the beam calibration. Subsequent follow-up actions help to resolve the discrepancies, thus preventing further mistreatment of patients. The audits for SSDLs check the implementation of the dosimetry protocol in order to assure proper dissemination of dosimetry standards to the end-users. The TLD audit results for SSDLs show good consistency in the basic dosimetry worldwide. New TLD procedures and equipment have recently been introduced by the IAEA that include a modified TLD calibration methodology and computerised tools for automation of dose calculation from TLD readings. (author)

  2. Optimisation of an EPR dosimetry system for robust and high precision dosimetry

    International Nuclear Information System (INIS)

    Clinical applications of electron paramagnetic resonance (EPR) dosimetry systems demand high accuracy causing time consuming analysis. The need for high spatial resolution dose measurements in regions with steep dose gradients demands small sized dosimeters. An optimization of the analysis was therefore needed to limit the time consumption. The aim of this work was to introduce a new smaller lithium formate dosimeter model (diameter reduced from standard diameter 4.5 mm to 3 mm and height from 4.8 mm to 3 mm). To compensate for reduced homogeneity in a batch of the smaller dosimeters, a method for individual sensitivity correction suitable for EPR dosimetry was tested. Sensitivity and repeatability was also tested for a standard EPR resonator and a super high Q (SHQE) one. The aim was also to optimize the performance of the dosimetry system for better efficiency regarding measurement time and precision. A systematic investigation of the relationship between measurement uncertainty and number of readouts per dosimeter was performed. The conclusions drawn from this work were that it is possible to decrease the dosimeter size with maintained measurement precision by using the SHQE resonator and introducing individual calibration factors for dosimeter batches. It was also shown that it is possible reduce the number of readouts per dosimeter without significantly decreasing the accuracy in measurements. - Highlights: • A lithium formate dosimetry system was optimized for accurate dose determinations. • Smaller-sized dosimeters for measurements in dose gradient regions was developed. • Individual sensitivity calibration was introduced for EPR dosimetry. • Measurement precision versus measurement time was evaluated

  3. A STUDY ON THE UNCERTAINTY FOR THE ROUTINE DOSIMETRY SERVICE AT THE LEBANESE ATOMIC ENERGY COMMISSION USING HARSHAW 8814 DOSEMETERS.

    Science.gov (United States)

    Rizk, C; Vanhavere, F

    2016-09-01

    The personal dosimetry service at the Lebanese Atomic Energy Commission uses Harshaw 8814 cards with LiF:Mg,Ti detectors. The dosemeters are read in a Harshaw 6600 TLD reader. In the process of accreditation for the ISO 17025 standard((1)), different influence factors are investigated and the uncertainty has been determined. The Individual Monitoring Service Laboratory-LAEC reads the dosemeters once it receives them from the customer, and new cards are immediately given for the next wearing period. The wearing period is 2 months. The dosemeter results are reported to the customers without background subtraction. Both Hp(10) and Hp(0.07) are reported. For this paper, only the uncertainty on Hp(10) will be focussed. The following factors are taken into account for the uncertainty: calibration factor, dosemeter homogeneity and repeatability, energy and angular dependence, non-linearity, temperature dependence, etc. Also the detection limit was determined. One of the important factors is the correction for fading. This fading correction depends on the procedure used such as storage temperatures, the time-temperature profile of the read-out, pre-heat and annealing conditions. Pre- and post-irradiation fading curves were measured for a storage period up to 182 d at room temperature (15-25°C). The resulting final combined standard uncertainty on the reported doses is of the order of 24 % for doses of ∼1 mSv. PMID:26443544

  4. Accreditation to manage research programs

    International Nuclear Information System (INIS)

    In this report for an accreditation to supervise research, the author proposes an overview of a study of transfers of vanadium towards benthic organisms (i.e. the toxicity of vanadium for sea coastal organisms), of studies of transfer of transuranic elements from sediment to marine benthic species. He presents current researches and perspectives: study of the level of metallic pollutants and physical-chemical characteristics of coastal waters in northern Cotentin, researches in Seine Bay, study of pollution biologic indicators. Numerous articles are provided in appendix

  5. Accredited Internship and Postdoctoral Programs for Training in Psychology: 2012

    Science.gov (United States)

    American Psychologist, 2012

    2012-01-01

    This is the official listing of accredited internship and postdoctoral residency programs in psychology. It reflects all Commission on Accreditation decisions through July 22, 2012. (Contains 15 footnotes.)

  6. The Accreditation Process for Engineering Curricula with Cooperative Education.

    Science.gov (United States)

    Bostwick, Williard D.

    1980-01-01

    Details the structure, accreditation procedures, and criteria of the Accreditation Board for Engineering and Technology, with particular attention to the assessment of cooperative engineering education programs. (SK)

  7. Dosimetry quality assurance in Martin Marietta Energy Systems' centralized external dosimetry system

    International Nuclear Information System (INIS)

    External dosimetry needs at the four Martin Marietta Energy Systems facilities are served by Energy Systems Centralized External Dosimetry System (CEDS). The CEDS is a four plant program with four dosimeter distribution centers and two dosimeter processing centers. Each plant has its own distribution center, while processing centers are located at ORNL and the Y-12 Plant. The program has been granted accreditation by the Department of Energy Laboratory Accreditation Program (DOELAP). The CEDS is a TLD based system which is responsible for whole-body beta-gamma, neutron, and extremity monitoring. Beta-gamma monitoring is performed using the Harshaw/Solon Technologies model 8805 dosimeter. Effective October 1, 1992 the standard silver mylar has been replaced with an Avery mylar foil blackened on the underside with ink. This was done in an effort to reduce the number of light induced suspect readings. At this time we have little operational experience with the new blackened mylars-The CEDS neutron dosimeter is the Harshaw model 8806B. This card/holder configuration contains two TLD-600/TLD-700 chip pairs; one pair is located beneath a cadmium filter and one pair is located beneath a plastic filter. In routine personnel monitoring the CEDS neutron dosimeter is always paired with a CEDS beta-gamma dosimeter.The CEDS extremity dosimeter is composed of a Harshaw thin TLD-700 dosiclip placed inside a Teledyne RB-4 finger sachet. The finger sachet provides approximately 7 mg/cm2 filtration over the chip. A teflon ring surrounds the dosiclip to help prevent tearing of the vinyl sachet

  8. Comments on "Reinventing Social Work Accreditation"

    Science.gov (United States)

    Midgley, James

    2009-01-01

    It is unlikely that Stoesz and Karger will be widely commended for the critique of social work accreditation. Social work academics do not usually handle criticism with equanimity. In some respects, their case is overstated. The problems associated with social work accreditation are not caused by the low publication productivity of social work…

  9. Social Partnership in Accrediting Lithuanian VET Qualifications

    Science.gov (United States)

    Tutlys, Vidmantas; Kaminskiene, Lina

    2008-01-01

    This article examines social partnership in accrediting qualifications in Lithuania. It defines the factors influencing social partnership and surveys future development perspectives, referring to the creation and implementation of the national qualifications system in Lithuania. Social partnership in qualifications accreditation is regarded as a…

  10. Accreditation and Expansion in Danish Higher Education

    DEFF Research Database (Denmark)

    Rasmussen, Palle

    2014-01-01

    During the last decade, an accreditation system for higher education has been introduced in Denmark. Accreditation partly represents continuity from an earlier evaluation system, but it is also part of a government policy to increasingly define higher education institutions as market actors. The ...

  11. ACCREDITATION FOR TECHNICAL ABILITIES INCLUDING COMPUTER SKILLS

    Directory of Open Access Journals (Sweden)

    Halit Hami OZ

    2013-01-01

    Full Text Available Sector Skills are defined by state-sponsored, employer-led organizations that cover specific economic sectors in the European Union and other countries in the world to reduce skills gaps and shortages, improve productivity, boost the skills of their sector workforces and improve learning supply. The accreditation and registration systems used by professional bodies raise the profile of the profession. In many countries including the European Union, professional associations are beginning to accept practice-based accreditation, generally as an alternative to their mainstream systems. Besides studying the certain agencies in the European Union for assessing/accreditating practical abilities , Accreditation for practical abilities of Information Communication Technology and Business Management/Language domains developed by Accreditation Council for Practical abilities are also studied in detail as an example to establish a similar agency in Turkey.

  12. A Finnish national code of practice for reference dosimetry of radiation therapy

    International Nuclear Information System (INIS)

    Full text: A national Code of Practice (CoP) for reference dosimetry of radiation therapy in Finland will be established during 2002 and will be implemented from the beginning of 2003. The CoP will cover dosimetry of the conventional radiotherapy modalities used in Finland i.e. external radiotherapy with megavoltage photon and electron beams, external radiotherapy with low energy kilovoltage X-ray beams and brachytherapy. The formalisms for external radiation beam dosimetry are those of TRS 389. For brachytherapy the formalism will follow the general guidelines of TECDOC-1274. The CoP will be prepared by the SSDL of STUK in close co-operation with the Finnish radiotherapy physicists. For external beam radiotherapy, the main objective of the national Code of Practice for radiation therapy dosimetry is to maintain the achieved good level of consistency of the dosimetry procedures in external beam radiotherapy as the 'absorbed dose to water' based approach of TRS 389 is implemented in Finland. In the CoP the dosimetry the procedures are described for the whole dosimetry chain starting from the calibration of the ionisation chambers at the SSDL of STUK and ending to the calibration of the beam monitor ionisation chamber of a linear accelerator. For brachytherapy dosimetry the aim is to fix the national practice for reference air kerma rate calibrations both for radioactive sources and for well-type ionisation chambers. Although the dosimetry procedures are described independently of the SSDL service, CoP makes use of the special features of the calibration service offered by the SSDL of STUK. For ionisation chambers used for photon dosimetry the calibration factors for the user measurement chain are given not only for the actual reference beam quality (60Co) but also for a set of user beam qualities. Furthermore, SSDL of STUK offers calibration services for plane parallel ionisation chambers in an electron beam of a user linac. For brachytherapy SSDL of STUK has

  13. In vivo dosimetry with silicon diodes in total body irradiation

    Science.gov (United States)

    Oliveira, F. F.; Amaral, L. L.; Costa, A. M.; Netto, T. G.

    2014-02-01

    The aim of this work is the characterization and application of silicon diode detectors for in vivo dosimetry in total body irradiation (TBI) treatments. It was evaluated the diode response with temperature, dose rate, gantry angulations and field size. A maximum response variation of 2.2% was obtained for temperature dependence. The response variation for dose rate and angular was within 1.2%. For field size dependence, the detector response increased with field until reach a saturation region, where no more primary radiation beam contributes for dose. The calibration was performed in a TBI setup. Different lateral thicknesses from one patient were simulated and then the calibration factors were determined by means of maximum depth dose readings. Subsequent to calibration, in vivo dosimetry measurements were performed. The response difference between diode readings and the prescribed dose for all treatments was below 4%. This difference is in agreement as recommended by the International Commission on Radiation Units and Measurements (ICRU), which is ±5%. The present work to test the applicability of a silicon diode dosimetry system for performing in vivo dose measurements in TBI techniques presented good results. These measurements demonstrated the value of diode dosimetry as a treatment verification method and its applicability as a part of a quality assurance program in TBI treatments.

  14. The current status of forensic science laboratory accreditation in Europe.

    Science.gov (United States)

    Malkoc, Ekrem; Neuteboom, Wim

    2007-04-11

    Forensic science is gaining some solid ground in the area of effective crime prevention, especially in the areas where more sophisticated use of available technology is prevalent. All it takes is high-level cooperation among nations that can help them deal with criminality that adopts a cross-border nature more and more. It is apparent that cooperation will not be enough on its own and this development will require a network of qualified forensic laboratories spread over Europe. It is argued in this paper that forensic science laboratories play an important role in the fight against crime. Another, complimentary argument is that forensic science laboratories need to be better involved in the fight against crime. For this to be achieved, a good level of cooperation should be established and maintained. It is also noted that harmonization is required for such cooperation and seeking accreditation according to an internationally acceptable standard, such as ISO/IEC 17025, will eventually bring harmonization as an end result. Because, ISO/IEC 17025 as an international standard, has been a tool that helps forensic science laboratories in the current trend towards accreditation that can be observed not only in Europe, but also in the rest of the world of forensic science. In the introduction part, ISO/IEC 17025 states that "the acceptance of testing and calibration results between countries should be facilitated if laboratories comply with this international standard and if they obtain accreditation from bodies which have entered into mutual recognition agreements with equivalent bodies in other countries using this international standard." Furthermore, it is emphasized that the use of this international standard will assist in the harmonization of standards and procedures. The background of forensic science cooperation in Europe will be explained by using an existing European forensic science network, i.e. ENFSI, in order to understand the current status of forensic

  15. Nuclear medicine radiation dosimetry

    CERN Document Server

    McParland, Brian J

    2010-01-01

    Complexities of the requirements for accurate radiation dosimetry evaluation in both diagnostic and therapeutic nuclear medicine (including PET) have grown over the past decade. This is due primarily to four factors: growing consideration of accurate patient-specific treatment planning for radionuclide therapy as a means of improving the therapeutic benefit, development of more realistic anthropomorphic phantoms and their use in estimating radiation transport and dosimetry in patients, design and use of advanced Monte Carlo algorithms in calculating the above-mentioned radiation transport and

  16. INDIVIDUAL DOSIMETRY SERVICE

    CERN Multimedia

    2000-01-01

    Personnel in the distribution groups Aleph, Delphi, L3, Opal who also work for other experiments than at LEP, should contact their dispatchers to explain their activities for the future, after LEP dismantling in order to be maintained on the regular distribution list at Individual DosimetryWe inform all staff and users under regular dosimetric control that the dosimeters for the monitoring period MAY/JUNE will be available from their usual dispatchers on Tuesday 2 May.Please have your films changed before the 12 May.The colour of the dosimeter valid in is MAY/JUNE is YELLOW.Individual Dosimetry Service will be closed on Friday 28 April.

  17. A new method for dosimetry with films radiochromic; Un nuevo metodo para la dosimetria con peliculas radiocromica

    Energy Technology Data Exchange (ETDEWEB)

    Mendez Carot, I.

    2013-07-01

    in this paper a new method is presented and the results of the comparison between the calibration is summarized based on a planning reference and calibration obtained from the irradiated fragments measure different dose levels multichannel compare dosimetry based on the weighted average dosimetry described by Micke et al.(present in the FilmQAPro software) and, finally, show different results obtained with the method proposed in several applications clinics. (Author)

  18. Status of radiation processing dosimetry

    DEFF Research Database (Denmark)

    Miller, A.

    1993-01-01

    Several milestones have marked the field of radiation processing dosimetry since IMRP 7. Among them are the IAEA symposium on High Dose Dosimetry for Radiation Processing and the international Workshops on Dosimetry for Radiation Processing organized by the ASTM. Several standards have been...... or are being published by the ASTM in this field, both on dosimetry procedures and on the proper use of specific dosimeter systems. Several individuals are involved in this international cooperation which contribute significantly to the broader understanding of the role of dosimetry in radiation processing....... The importance of dosimetry is emphasized in the standards on radiation sterilization which are currently drafted by the European standards organization CEN and by the international standards organization ISO. In both standards, dosimetry plays key roles in characterization of the facility, in qualification...

  19. Accreditation - Its relevance for laboratories measuring radionuclides

    Energy Technology Data Exchange (ETDEWEB)

    Palsson, S.E. [Icelandic Radiation Protection Inst. (Iceland)

    2001-11-01

    Accreditation is an internationally recognised way for laboratories to demonstrate their competence. Obtaining and maintaining accreditation is, however, a costly and time-consuming procedure. The benefits of accreditation also depend on the role of the laboratory. Accreditation may be of limited relevance for a research laboratory, but essential for a laboratory associated with a national authority and e.g. issuing certificates. This report describes work done within the NKSBOK-1.1 sub-project on introducing accreditation to Nordic laboratories measuring radionuclides. Initially the focus was on the new standard ISO/IEC 17025, which was just in a draft form at the time, but which provides now a new framework for accreditation of laboratories. Later the focus was widened to include a general introduction to accreditation and providing through seminars a forum for exchanging views on the experience laboratories have had in this field. Copies of overheads from the last such seminar are included in the appendix to this report. (au)

  20. 21 CFR 900.4 - Standards for accreditation bodies.

    Science.gov (United States)

    2010-04-01

    ... image quality, or upon request by FDA, the accreditation body shall review a facility's clinical images... review by the accreditation body demonstrates that a problem does exist with respect to image quality or... accreditation body shall review clinical images from each facility accredited by the body at least once every...

  1. Ion storage dosimetry

    Science.gov (United States)

    Mathur, V. K.

    2001-09-01

    The availability of a reliable, accurate and cost-effective real-time personnel dosimetry system is fascinating to radiation workers. Electronic dosimeters are contemplated to meet this demand of active dosimetry. The development of direct ion storage (DIS) dosimeters, a member of the electronic dosimeter family, for personnel dosimetry is also an attempt in this direction. DIS dosimeter is a hybrid of the well-established technology of ion chambers and the latest advances in data storage using metal oxide semiconductor field effect transistor (MOSFET) analog memory device. This dosimeter is capable of monitoring legal occupational radiation doses of gamma, X-rays, beta and neutron radiation. Similar to an ion chamber, the performance of the dosimeter for a particular application can be optimized through the selection of appropriate wall materials. The use of the floating gate of a MOSFET as one of the electrodes of the ion chamber allows the miniaturization of the device to the size of a dosimetry badge and avoids the use of power supplies during dose accumulation. The concept of the device, underlying physics and the design of the DIS dosimeter are discussed. The results of preliminary testing of the device are also provided.

  2. High frequency electromagnetic dosimetry

    CERN Document Server

    Sánchez-Hernández, David A

    2009-01-01

    Along with the growth of RF and microwave technology applications, there is a mounting concern about the possible adverse effects over human health from electromagnetic radiation. Addressing this issue and putting it into perspective, this groundbreaking resource provides critical details on the latest advances in high frequency electromagnetic dosimetry.

  3. Intercalibration of radiological measurements for surveillance purposes of the internal dosimetry laboratory coordinated by the IAEA

    International Nuclear Information System (INIS)

    The ININ of Mexico participated in this intercomparison organized by the IAEA in 2000. The objective of this activity is that the dosimetry laboratories that participate can validate the programs of internal dosimetry, with the purpose of improving its capacity in the evaluation of the internal dose and have access to a mechanism to evaluate its dosimetry system under real conditions. The specific objectives of this intercomparison were: 1. To evaluate the participant's capacity to manage the measurements of individual monitoring in terms of the activity in the phantom. 2. To provide the access to the unique calibration resources that otherwise would not be available. 3. To compare the operation of several detection systems, the geometry, phantoms, calibration methods and methods for the evaluation of activity of the radionuclide used by each institution. 4. To provide the independent verification of the direct measurement methods of the dosimetry service. (Author)

  4. Development of Kilovoltage X-ray Dosimetry Methods and Their Application to Cone Beam Computed Tomography

    Science.gov (United States)

    Lawless, Michael J.

    The increase in popularity of pre-treatment imaging procedures in radiation therapy, such as kilovoltage cone beam computed tomography (CBCT), has been accompanied by an increase in the dose delivered to the patient from these imaging procedures. The measurement of dose from CBCT scans is complicated, as currently available kilovoltage dosimetry protocols are based on air-kerma standards and radiation detectors exhibit large energy responses at the low photon energies used in the imaging procedures. This work aims to provide the tools and methodology needed to measure the dose from these scans more accurately and precisely. Through the use of a validated Monte Carlo (MC) model of the moderately filtered (M-series) x-ray beams at the University of Wisconsin Accredited Dosimetry Calibration Laboratory, dose-to-water rates were obtained in a water phantom for the M-series x-ray beams with tube potentials from 40-250 kVp. The resulting dose-to-water rates were consistent with previously established methods, but had significantly reduced uncertainties. While detectors are commonly used to measure dose in phantom, previous investigations of the energy response of common detectors in the kilovoltage energy range have been limited to in-air geometries. The newly determined dose-to-water rates were used to characterize the in-phantom energy and depth response of thermoluminescent dosimeters and ionization chambers. When compared to previous investigations of the in-air detector response, the impact of scatter and absorption of the photon beam by the water medium was found to have a significant impact on the response of certain detectors. The dose to water in the NIST-traceable M-series x-ray beams was transferred to clinical CBCT beams and the resulting doses agreed with other dose-to-water measurement techniques. The dose to water in the CBCT beams was used to characterize the energy and depth responses of a number of detectors. The energy response in the CBCT beams agreed

  5. Dosimetry and process control for food irradiation

    International Nuclear Information System (INIS)

    Whatever a radiation process is designed to achieve, dosimetry is fundamental to it, either as a necessary control, or to establish the process, or for research and development studies. Dosimetry provides the quantitative baseline against which the biological or chemical changes induced by radiation can be measured. In the case of irradiation of food, a minimum dose will be required to achieve the technological objective. A maximum dose will be defined by the onset of the loss of acceptable quality of the food, but upper dose limits will usually be prescribed by regulatory bodies. There is no accurate way of assessing the dose once the food has left the irradiation plant. Therefore the dose must be properly applied and verified during processing. The dose is measured using dosimeters. There are many different types of dosimeters for different applications, dose ranges and conditions of use. All dosimeters must be calibrated, with a measurement traceability chain to a national or international primary standard. This paper describes the classification of dosimeters and gives examples and their applications. Calibration and use of a typical dosimetry system used for food irradiation is then discussed, including the effects of environmental influence factors such as dose rate and temperature, and how measurement traceability can be established. Before routine processing of a product can occur, process qualification must be carried out to ensure the irradiation process produces acceptable results. An example of a dose mapping study is given, followed by discussion of some practical considerations of process control, including measurement uncertainty and how this relates to the setting of process limits

  6. Verification of IMRT fields by film dosimetry

    International Nuclear Information System (INIS)

    In intensity modulated radiation therapy (IMRT) the aim of an accurate conformal dose distribution is obtained through a complex process. This ranges from the calculation of the optimal distribution of fluence by the treatment planning system (TPS), to the dose delivery through a multilamellar collimator (MLC), with several segments per beam in the step and shoot approach. The above-mentioned consideration makes mandatory an accurate dosimetric verification of the IM beams. A high resolution and integrating dosimeter, like the radiographic film, permits one to simultaneously measure the dose in a matrix of points, providing a good means of obtaining dose distributions. The intrinsic limitation of film dosimetry is the sensitivity dependence on the field size and on the measurement depth. However, the introduction of a scattered radiation filter permits the use of a single calibration curve for all field sizes and measurement depths. In this paper the quality control procedure developed for dosimetric verification of IMRT technique is reported. In particular a system of film dosimetry for the verification of a 6 MV photon beam has been implemented, with the introduction of the scattered radiation filter in the clinical practice that permits one to achieve an absolute dose determination with a global uncertainty within 3.4% (1 s.d.). The film has been calibrated to be used both in perpendicular and parallel configurations. The work also includes the characterization of the Elekta MLC. Ionimetric independent detectors have been used to check single point doses. The film dosimetry procedure has been applied to compare the measured absolute dose distributions with the ones calculated by the TPS, both for test and clinical plans. The agreement, quantified by the gamma index that seldom reaches the 1.5 value, is satisfying considering that the comparison is performed between absolute doses

  7. Neutron dosimetry; Dosimetria de neutrons

    Energy Technology Data Exchange (ETDEWEB)

    Fratin, Luciano

    1993-12-31

    A neutron irradiation facility was designed and built in order to establish a procedure for calibrating neutron monitors and dosemeters. A 185 GBq {sup 241} Am Be source of known is used as a reference source. The irradiation facility using this source in the air provides neutron dose rates between 9 nSv s{sup -1} and 0,5 {sup {mu}}Sv s{sup -1}. A calibrated 50 nSv s{sup -1} thermal neutron field is obtained by using a specially designed paraffin block in conjunction with the {sup 241} Am Be source. A Bonner multisphere spectrometer was calibrated, using a procedure based on three methods proposed by international standards. The unfold {sup 241} Am Be neutron spectrum was determined from the Bonner spheres data and resulted in a good agreement with expected values for fluence rate, dose rate and mean energy. A dosimetric system based on the electrochemical etching of CR-39 was developed for personal dosimetry. The dosemeter badge using a (n,{alpha}) converter, the etching chamber and high frequency power supply were designed and built specially for this project. The electrochemical etching (ECE) parameters used were: a 6N KOH solution, 59 deg C, 20 kV{sub pp} cm{sup -1}, 2,0 kHz, 3 hours of ECE for thermal and intermediate neutrons and 6 hours for fast neutrons. The calibration factors for thermal, intermediate and fast neutrons were determined for this personal dosemeter. The sensitivities determined for the developed dosimetric system were (1,46{+-} 0,09) 10{sup 4} tracks cm{sup -2} mSv{sup -1} for thermal neutrons, (9{+-}3) 10{sup 2} tracks cm{sup -2} mSV{sup -1} for intermediate neutrons and (26{+-}4) tracks cm{sup -2} mSv{sup -1} for fast neutrons. The lower and upper limits of detection were respectively 0,002 mSv and 0,6 mSv for thermal neutrons, 0,04 mSv and 8 mSv for intermediate neutrons and 1 mSv and 12 mSv for fast neutrons. In view of the 1990`s ICRP recommendations, it is possible to conclude that the personal dosemeter described in this work is

  8. On multichannel film dosimetry with channel-independent perturbations

    International Nuclear Information System (INIS)

    Purpose: Different multichannel methods for film dosimetry have been proposed in the literature. Two of them are the weighted mean method and the method put forth byMicke et al. [“Multichannel film dosimetry with nonuniformity correction,” Med. Phys. 38, 2523–2534 (2011)] and Mayer et al. [“Enhanced dosimetry procedures and assessment for EBT2 radiochromic film,” Med. Phys. 39, 2147–2155 (2012)]. The purpose of this work was to compare their results and to develop a generalized channel-independent perturbations framework in which both methods enter as special cases. Methods: Four models of channel-independent perturbations were compared: weighted mean, Micke–Mayer method, uniform distribution, and truncated normal distribution. A closed-form formula to calculate film doses and the associated type B uncertainty for all four models was deduced. To evaluate the models, film dose distributions were compared with planned and measured dose distributions. At the same time, several elements of the dosimetry process were compared: film type EBT2 versus EBT3, different waiting-time windows, reflection mode versus transmission mode scanning, and planned versus measured dose distribution for film calibration and for γ-index analysis. The methods and the models described in this study are publicly accessible through IRISEU. Alpha 1.1 ( http://www.iriseu.com ). IRISEU. is a cloud computing web application for calibration and dosimetry of radiochromic films. Results: The truncated normal distribution model provided the best agreement between film and reference doses, both for calibration and γ-index verification, and proved itself superior to both the weighted mean model, which neglects correlations between the channels, and the Micke–Mayer model, whose accuracy depends on the properties of the sensitometric curves. With respect to the selection of dosimetry protocol, no significant differences were found between transmission and reflection mode scanning

  9. Validation protocol for multiple blood gas analyzers in accordance with laboratory accreditation programs

    OpenAIRE

    Pérsio A. R. Ebner; Paschoalina Romano; Alexandre Sant’Anna; Maria Elizabete Mendes; Magna Oliveira; Nairo M. Sumita

    2015-01-01

    ABSTRACTIntroduction:The results of blood gas analysis using different instrumentation can vary widely due to the methodological differences, the calibration procedures and the use of different configurations for each type of instrument.Objective:The objective of this study was to evaluate multiple analytical systems for measurement of blood gases, electrolytes and metabolites in accordance with the accreditation program (PALC) of Sociedade Brasileira de Patologia Clínica/Medicina Laboratoria...

  10. Twenty new ISO standards on dosimetry for radiation processing

    International Nuclear Information System (INIS)

    Twenty standards on essentially all aspects of dosimetry for radiation processing were published as new ISO standards in December 1998. The standards are based on 20 standard practices and guides developed over the past 14 years by Subcommittee E10.01 of the American Society for Testing and Materials (ASTM). The transformation to ISO standards using the 'fast track' process under ISO Technical Committee 85 (ISO/TC85) commenced in 1995 and resulted in some overlap of technical information between three of the new standards and the existing ISO Standard 11137 Sterilization of health care products - Requirements for validation and routine control - Radiation sterilization. Although the technical information in these four standards was consistent, compromise wording in the scopes of the three new ISO standards to establish precedence for use were adopted. Two of the new ISO standards are specifically for food irradiation applications, but the majority apply to all forms of gamma, X-ray, and electron beam radiation processing, including dosimetry for sterilization of health care products and the radiation processing of fruit, vegetables, meats, spices, processed foods, plastics, inks, medical wastes, and paper. Most of the standards provide exact procedures for using individual dosimetry systems or for characterizing various types of irradiation facilities, but one covers the selection and calibration of dosimetry systems, and another covers the treatment of uncertainties using the new ISO Type A and Type B evaluations. Unfortunately, nine of the 20 standards just adopted by the ISO are not the most recent versions of these standards and are therefore already out of date. To help solve this problem, efforts are being made to develop procedures to coordinate the ASTM and ISO development and revision processes for these and future ASTM-originating dosimetry standards. In the meantime, an additional four dosimetry standards have recently been published by the ASTM but have

  11. Accreditation standards for undergraduate forensic science programs

    Science.gov (United States)

    Miller, Marilyn Tebbs

    Undergraduate forensic science programs are experiencing unprecedented growth in numbers of programs offered and, as a result, student enrollments are increasing. Currently, however, these programs are not subject to professional specialized accreditation. This study sought to identify desirable student outcome measures for undergraduate forensic science programs that should be incorporated into such an accreditation process. To determine desirable student outcomes, three types of data were collected and analyzed. All the existing undergraduate forensic science programs in the United States were examined with regard to the input measures of degree requirements and curriculum content, and for the output measures of mission statements and student competencies. Accreditation procedures and guidelines for three other science-based disciplines, computer science, dietetics, and nursing, were examined to provide guidance on accreditation processes for forensic science education programs. Expert opinion on outcomes for program graduates was solicited from the major stakeholders of undergraduate forensic science programs-forensic science educators, crime laboratory directors, and recent graduates. Opinions were gathered by using a structured Internet-based survey; the total response rate was 48%. Examination of the existing undergraduate forensic science programs revealed that these programs do not use outcome measures. Of the accreditation processes for other science-based programs, nursing education provided the best model for forensic science education, due primarily to the balance between the generality and the specificity of the outcome measures. From the analysis of the questionnaire data, preliminary student outcomes, both general and discipline-specific, suitable for use in the accreditation of undergraduate forensic science programs were determined. The preliminary results were reviewed by a panel of experts and, based on their recommendations, the outcomes

  12. A new implementation of multichannel radiochromic film dosimetry

    International Nuclear Information System (INIS)

    The aims of this paper are to carry out a new implementation of the multichannel radiochromic film dosimetry (Micke A, Lewis D, Yu X. Multichannel film dosimetry with nonuniformity correction. Med Phys 2011;38:2523-34), to quantify the variation in gamma index as compared to the single channel film dosimetry, and to determine if the procedure achieves similar results by means of a different scanner that the one used by Micke et al. Radiochromic EBT2 films and a Microtek 9000 XL scanner were used. Our procedure simplifies the system calibration splitting it into two factors, manufactured batch and digitalization specific. Absorbed dose spatial distributions from an open radiotherapy beam without any modulation and 20 IMRT treatments were determined. Their gamma index maps were calculated and a comparison of the results from single channel and multichannel dosimetry was performed. A 5% mean increase in concordance was obtained by using the multichannel film dosimetry. Our results are similar to those reported by Micke et al. even though we are using a different scanner. (Author)

  13. In vivo dosimetry for IMRT

    Science.gov (United States)

    Vial, Philip

    2011-05-01

    In vivo dosimetry has a well established role in the quality assurance of 2D radiotherapy and 3D conformal radiotherapy. The role of in vivo dosimetry for IMRT is not as well established. IMRT introduces a range of technical issues that complicate in vivo dosimetry. The first decade or so of IMRT implementation has largely relied upon pre-treatment phantom based dose verification. During that time, several new devices and techniques for in vivo dosimetry have emerged with the promise of providing the ultimate form of IMRT dose verification. Solid state dosimeters continue to dominate the field of in vivo dosimetry in the IMRT era. In this report we review the literature on in vivo dosimetry for IMRT, with an emphasis on clinical evidence for different detector types. We describe the pros and cons of different detectors and techniques in the IMRT setting and the roles that they are likely to play in the future.

  14. The dosimetry of ionizing radiation

    CERN Document Server

    1990-01-01

    A continuation of the treatise The Dosimetry of Ionizing Radiation, Volume III builds upon the foundations of Volumes I and II and the tradition of the preceeding treatise Radiation Dosimetry. Volume III contains three comprehensive chapters on the applications of radiation dosimetry in particular research and medical settings, a chapter on unique and useful detectors, and two chapters on Monte Carlo techniques and their applications.

  15. Ambiguities in thermoluminescence dosimetry

    International Nuclear Information System (INIS)

    On one hand, thermoluminescence dosimetry is one of most reliable, rugged and economical system of passive dosimetry but on the other hand there are several ambiguities, which need attention. The PTTL is a complex phenomenon and it is difficult to identify the source for the transfer of the charge carrier to repopulate the traps related to the glow peaks. For the photon energy dependence it is difficult to explain the change in the response for 662 keV gamma rays of 137Cs as compared to the response for 1.25 MeV gamma rays of 60Co. The increase in the response of a TLD with increasing heating rate poses another ambiguity and so is the case with the observations of the supra linearity of different glow peaks. To over come the ambiguities, efforts have to continue to enhance the understanding and to harmonize the protocol for reliable experimental data

  16. Neutron beam measurement dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Amaro, C.R. [Idaho National Engineering Lab., Idaho Falls, ID (United States)

    1995-11-01

    This report describes animal dosimetry studies and phantom measurements. During 1994, 12 dogs were irradiated at BMRR as part of a 4 fraction dose tolerance study. The animals were first infused with BSH and irradiated daily for 4 consecutive days. BNL irradiated 2 beagles as part of their dose tolerance study using BPA fructose. In addition, a dog at WSU was irradiated at BMRR after an infusion of BPA fructose. During 1994, the INEL BNCT dosimetry team measured neutron flux and gamma dose profiles in two phantoms exposed to the epithermal neutron beam at the BMRR. These measurements were performed as a preparatory step to the commencement of human clinical trials in progress at the BMRR.

  17. Quantitative imaging for clinical dosimetry

    Science.gov (United States)

    Bardiès, Manuel; Flux, Glenn; Lassmann, Michael; Monsieurs, Myriam; Savolainen, Sauli; Strand, Sven-Erik

    2006-12-01

    Patient-specific dosimetry in nuclear medicine is now a legal requirement in many countries throughout the EU for targeted radionuclide therapy (TRT) applications. In order to achieve that goal, an increased level of accuracy in dosimetry procedures is needed. Current research in nuclear medicine dosimetry should not only aim at developing new methods to assess the delivered radiation absorbed dose at the patient level, but also to ensure that the proposed methods can be put into practice in a sufficient number of institutions. A unified dosimetry methodology is required for making clinical outcome comparisons possible.

  18. Relative dosimetry by Ebt-3; Dosimetria relativa por EBT3

    Energy Technology Data Exchange (ETDEWEB)

    De Leon A, M. A.; Rivera M, T. [IPN, Centro de Investigacion en Ciencia Aplicada y Tecnologia Avanzada, Av. Legaria 694, 11500 Mexico D. F. (Mexico); Hernandez O, J. O., E-mail: madla16@hotmail.com [Hospital General de Mexico, Dr. Balmis 148, Col. Doctores, 06726 Mexico D. F. (Mexico)

    2015-10-15

    In the present work relative dosimetry in two linear accelerator for radiation therapy was studied. Both Varian Oncology systems named Varian Clinac 2100-Cd and MLC Varian Clinac i X were used. Gaf Chromic Ebt-3 film was used. Measurements have been performed in a water equivalent phantom, using 6 MV and 18 MV photon beams on both Linacs. Both calibration and Electron irradiations were carried out with the ionization chamber placed at the isocenter, below a stack of solid water slabs, at the depth of dose maximum (D max), with a Source-to-Surface Distance (SSD) of 100 cm and a field size of 10 cm x 10 cm. Calibration and dosimetric measurements photons were carried out under IAEA-TRS 398 protocol. Results of relative dosimetry in the present work are discussed. (Author)

  19. Personnel radiation dosimetry

    International Nuclear Information System (INIS)

    The book contains the 21 technical papers presented at the Technical Committee Meeting to Elaborate Procedures and Data for the Intercomparison of Personnel Dosimeters organizaed by the IAEA on 22-26 April 1985. A separate abstract was prepared for each of these papers. A list of areas in which additional research and development work is needed and recommendations for an IAEA-sponsored intercomparison program on personnel dosimetry is also included

  20. INDIVIDUAL DOSIMETRY SERVICE

    CERN Multimedia

    1999-01-01

    Personnel in the distribution groups Aleph, Delphi, L3, Opal who also work for other experiments than at LEP, should contact the Individual Dosimetry ServiceWe inform all staff and users under regular dosimetric control that the dosimeters for the monitoring period JANUARY/FEBRUARY will be available from their usual dispatchers on Monday the third of January 2000.Please have your films changed:before the 12 January.The colour of the dosimeter valid in JANUARY/FEBRUARY is WHITE.

  1. INDIVIDUAL DOSIMETRY SERVICE

    CERN Multimedia

    2000-01-01

    Personnel in the distribution groups Aleph, Delphi, L3, Opal who also work for other experiments than at LEP, should contact the Individual Dosimetry Service.We inform all staff and users under regular dosimetric control that the dosimeters for the monitoring period MARCH/APRIL will be available from their usual dispatchers on the third of March 2000.Please have your films changed before the 13th of March.The colour of the dosimeter valid in MARCH/APRIL is BLUE.

  2. Personal radon daughter dosimetry

    International Nuclear Information System (INIS)

    The conventional means of radon daughter exposure estimatikn for uranium miners in Canada is by grab sampling and time weighting. Personal dosimetry is a possible alternative method with its own advantages and limitations. The author poses basic questions with regard to two methods of radon daughter detection, thermoluminescent chips and track-etch film. An historical review of previous and current research and development programs in Canada and in other countries is presented, as are brief results and conclusions of each dosimeter evaluation

  3. The UK radiotherapy dosimetry audit network

    International Nuclear Information System (INIS)

    Full text: Radiotherapy dosimetry intercomparison in the UK has been carried out in limited studies since the 1960s. However the first national dosimetry intercomparison involving all radiotherapy centres was conducted in the late 1980s. This was based on visits to each centre, using ionisation chamber dosimetry. It audited megavoltage photon beam calibration and other single field parameters. It also measured doses in a three-field 'treatment' in a trapezoidal phantom constructed from epoxy-resin water-equivalent material and compared these to locally planned doses. This included off-axis points, oblique incidence, inhomogeneities, etc. The study found mean measured beam calibration doses close to stated values (ratio 1.003), with a standard deviation (sd) of the distribution of 1.5% and 97% of doses within the pro-set 3% tolerance. For the planned multi-field irradiations, mean dose ratios (measured/stated) were 1.01 (sd 3%, 90% of results within 5%). A number of discrepancies were identified, leading to improved practice. A follow up study (mid-1990s) for electron beam audit also repeated the megavoltage photon calibration audit. For photons, an improvement was noted (mean ratio 1.003, sd 1.0%, 100% within 3%), whilst for electron beams, the mean ratio of measured/stated dose was 0.994 (sd 1.8%, 94% within 3%, 99% within 5%). In parallel with - and growing out of - this, a national audit network began to develop in 1991/2. It utilised similar methodology to the intercomparison and a network approach to allow parallel developments of the scope of the system. The network has eight regional groups, each with up to 10 radiotherapy centres, serving average populations of 7-8 million. Each group organises audits of its own centres and has developed at its own pace. Most have piloted methodology, phantoms, etc. for new audits which can then be used by other groups. All 65 UK centres are included. The network is co-ordinated by an IPEM Steering Committee (current chair

  4. IS 2010 and ABET Accreditation: An Analysis of ABET-Accredited Information Systems Programs

    Science.gov (United States)

    Saulnier, Bruce; White, Bruce

    2011-01-01

    Many strong forces are converging on information systems academic departments. Among these forces are quality considerations, accreditation, curriculum models, declining/steady student enrollments, and keeping current with respect to emerging technologies and trends. ABET, formerly the Accrediting Board for Engineering and Technology, is at…

  5. Proposed Accreditation Standards for Degree-Granting Correspondence Programs Offered by Accredited Institutions.

    Science.gov (United States)

    McGraw-Hill Continuing Education Center, Washington, DC.

    A study on proposed accreditation standards grew out of a need to (1) stimulate the growth of quality correspondence degree programs; and (2) provide a policy for accreditation of correspondence degree programs so that graduates would be encouraged to pursue advanced degree programs offered elsewhere by educational institutions. The study focused…

  6. Dosimetry: an ARDENT topic

    CERN Multimedia

    CERN Bulletin

    2012-01-01

    The first annual ARDENT workshop took place in Vienna from 20 to 23 November. The workshop gathered together the Early-Stage Researchers (ESR) and their supervisors, plus other people involved from all the participating institutions.   “The meeting, which was organised with the local support of the Austrian Institute of Technology, was a nice opportunity for the ESRs to get together, meet each other, and present their research plans and some preliminary results of their work,” says Marco Silari, a member of CERN Radiation Protection Group and the scientist in charge of the programme. Two full days were devoted to a training course on radiation dosimetry, delivered by renowned experts. The workshop closed with a half-day visit to the MedAustron facility in Wiener Neustadt. ARDENT (Advanced Radiation Dosimetry European Network Training) is a Marie Curie ITN project funded under EU FP7 with €4 million. The project focuses on radiation dosimetry exploiting se...

  7. Services of the CDRH X-ray calibration laboratory and their traceability to National Standards

    Energy Technology Data Exchange (ETDEWEB)

    Cerra, F.; Heaton, H.T. [Center for Devices and Radiological Health, Rockville, MD (United States)

    1993-12-31

    The X-ray Calibration Laboratory (XCL) of the Center for Devices and Radiological Health (CDRH) provides calibration services for the Food and Drug Administration (FDA). The instruments calibrated are used by FDA and contract state inspectors to verify compliance with federal x-ray performance standards and for national surveys of x-ray trends. In order to provide traceability of measurements, the CDRH XCL is accredited by the National Voluntary Laboratory Accreditation Program (NVLAP) for reference, diagnostic, and x-ray survey instrument calibrations. In addition to these accredited services, the CDRH XCL also calibrates non-invasive kVp meters in single- and three-phase x-ray beams, and thermoluminescent dosimeter (TLD) chips used to measure CT beam profiles. The poster illustrates these services and shows the traceability links back to the National Standards.

  8. US Navy's personnel neutron dosimetry program

    International Nuclear Information System (INIS)

    Since 1964 the Navy has been monitoring personnel for exposure to neutrons. With nuclear reactors, special nuclear materials, neutron calibration sources, and high energy linear accelerators, the Navy has organize over more neutron sources than any other single organization in the world. Personnel neutron monitoring in the Navy was first accomplished using NTA film from 1964 through 1980. In 1975 the Navy instituted albedo neutron dosimetry, which would completely replace NTA film in 1981. From 1975 through 1982 the Navy directed considerable efforts toward assessing the complexity of its field neutron spectra, evaluating techniques to correct the neutron response of its albedo dosimeter, and developing methods to perform neutron area monitoring. As a result of investigating field neutron spectra, the Navy again modified its calibration procedure in September 1979, by changing its calibration source to Cf-252 moderated with two inches of polyethylene. This source's spectrum more closely represented spectra found aboard ships, and as a result, shipboard personnel neutron doses were reduced by 45%. The Navy's newest albedo dosimeter is calibrated on an ANSI standard PMMA phantom using a D(sub 2)O moderated Cf-252 source at 50 cm

  9. Analysis of ISO/IEC 17025 for establishment of KOLAS (Korea Laboratory Accreditation Scheme) quality assurance system

    International Nuclear Information System (INIS)

    Besides one existent accredited lab, radioactive material chemical analysis lab, five test laboratories and two calibration labs are under plan to acquire the accreditation from KOLAS. But the current Quality Manual was developed according to ISO Guide 25 that was superceded by ISO/IEC 17025. Since it is tailored to the radioactive material chemical analysis lab, a number of requirements of the Manual are not applicable to the labs other than radioactive material chemical analysis lab. Through the analysis of ISO/IEC 17025, a model of quality system was established which is not only consistent with ISO/IEC 17025 but reflective of the KAERI's situation

  10. Analysis of ISO/IEC 17025 for establishment of KOLAS (Korea Laboratory Accreditation Scheme) quality assurance system

    Energy Technology Data Exchange (ETDEWEB)

    Nam, Ji Hee

    2000-12-01

    Besides one existent accredited lab, radioactive material chemical analysis lab, five test laboratories and two calibration labs are under plan to acquire the accreditation from KOLAS. But the current Quality Manual was developed according to ISO Guide 25 that was superceded by ISO/IEC 17025. Since it is tailored to the radioactive material chemical analysis lab, a number of requirements of the Manual are not applicable to the labs other than radioactive material chemical analysis lab. Through the analysis of ISO/IEC 17025, a model of quality system was established which is not only consistent with ISO/IEC 17025 but reflective of the KAERI's situation.

  11. The Radiotherapy Dosimetry Audit System In the UK

    International Nuclear Information System (INIS)

    Two national radiotherapy dosimetry intercomparisons have been earned out in the UK, involving all radiotherapy institutes. The first was concerned with megavoltage photon beams and looked at beam calibration and simple three-field planned distributions in a geometric phantom. The intercomparisons were carried out by an independent intercomparison physicist visiting each department in turn and making measurements with ion chambers, following a fixed protocol. The beam calibration intercomparison was earned out on every 60Co beam and every MV x-ray beam, whilst the planned comparisons were carried out on one beam only. The plans included effects of wedges, oblique incidence and inhomogeneities. The study was unfunded and took a significant time (1988-1991) to cover the 65 or so centres. It was followed up by a national electron dosimetry intercomparison which was fended (Department of Health) and which ran from 1994-1996. This audited three electron beam energies in each centre (depth dose, beam energy, dose calibration) and also included a follow-up of the original photon beam intercomparison. In general these studies showed good consistency of dosimetry across the UK centres, with mean (measured/locally stated) doses being close to unity and standard deviations of the distributions of values being approx. 1.5 and 1% for photons, 1.8% for electrons for beam calibration and 2.5-3.5% for the planned multi-beam situations. 97-100% of measurements were within the pre-set 3% tolerance for beam calibration and around 90% of the measurements within a pre-set 5% tolerance for planned situations. The studies did highlight some areas where increased on Q A could provide benefits. In particular the photon intercomparison discovered one 60Co unit mis calibration which led to national recommendations for the implementation of Quality Systems in radiotherapy departments

  12. Performance of a parallel plate ionization chamber in beta radiation dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Antonio, Patricia L.; Caldas, Linda V.E., E-mail: patrilan@ipen.b, E-mail: lcaldas@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)

    2011-07-01

    A homemade parallel plate ionization chamber with graphite collecting electrode, and developed for use in mammography beams, was tested in relation to its usefulness in beta radiation dosimetry at the Calibration Laboratory of IPEN. Characterization tests of this ionization chamber were performed, using the Sr-90 + Y-90, Kr-85 and Pm-147 sources of a beta secondary standard system. The results of saturation, leakage current, stabilization time, response stability, linearity, angular dependence, and calibration coefficients are within the recommended limits of international recommendations that indicate that this chamber may be used for beta radiation dosimetry. (author)

  13. Hanford External Dosimetry Technical Basis Manual PNL-MA-842

    Energy Technology Data Exchange (ETDEWEB)

    Rathbone, Bruce A.

    2007-03-12

    The Hanford External Dosimetry Technical Basis Manual PNL-MA-842 documents the design and implementation of the external dosimetry system used at Hanford. The manual describes the dosimeter design, processing protocols, dose calculation methodology, radiation fields encountered, dosimeter response characteristics, limitations of dosimeter design under field conditions, and makes recommendations for effective use of the dosimeters in the field. The manual describes the technical basis for the dosimetry system in a manner intended to help ensure defensibility of the dose of record at Hanford and to demonstrate compliance with 10 CFR 835, DOELAP, DOE-RL, ORP, PNSO, and Hanford contractor requirements. The dosimetry system is operated by PNNL’s Hanford External Dosimetry Program (HEDP) which provides dosimetry services to all Hanford contractors. The primary users of this manual are DOE and DOE contractors at Hanford using the dosimetry services of PNNL. Development and maintenance of this manual is funded directly by DOE and DOE contractors. Its contents have been reviewed and approved by DOE and DOE contractors at Hanford through the Hanford Personnel Dosimetry Advisory Committee (HPDAC) which is chartered and chaired by DOE-RL and serves as means of coordinating dosimetry practices across contractors at Hanford. This manual was established in 1996. Since inception, it has been revised many times and maintained by PNNL as a controlled document with controlled distribution. Rev. 0 marks the first revision to be released through PNNL’s Electronic Records & Information Capture Architecture (ERICA) database. Revision numbers that are whole numbers reflect major revisions typically involving changes to all chapters in the document. Revision numbers that include a decimal fraction reflect minor revisions, usually restricted to selected chapters or selected pages in the document. Revision Log: Rev. 0 (2/25/2005) Major revision and expansion. Rev. 0.1 (3/12/2007) Minor

  14. Proficiency test in the accreditation system

    Science.gov (United States)

    Legarda, F.; Herranz, M.; Idoeta, R.

    2008-08-01

    In the accreditation process of a radioactivity measurements laboratory, according to ISO standard 17025, proficiency tests play a fundamental role. These PTs constitute an irreplaceable tool for the validation of measuring methods. In the case of Spain, ENAC, which is the Spanish accreditation national body, requires that the laboratory has to take part in a PT for each one of the accredited measuring methods in the period of time between two reassessments of the accreditation, what happens every 4-5 years. In specific areas of determination procedures, among which radioactive measurements could be included, the number of methods which can be accredited is very large. The purpose of the present work is to establish a classification into families of the different radioactivity measurement procedures, as well as to establish complementary actions that guarantee that carrying out periodically proficiency-tests on any of the included procedures in each family, every measurement procedure include in that family is controlled, complying with the criteria established by ENAC.

  15. Answer to request on the ININ internal dosimetry

    International Nuclear Information System (INIS)

    In this report it is presented the reply to CNSNS asking for information about the methodology for the evaluation of the occupational dose due to internal contamination. The characteristics of the installation, type and dimensions of the shield room, construction materials, type of detecting, calibration geometries, type of used phantom, intervals of energy of the calibration, type of routine measurements, detection limit for Cs-137 and Co-60, code to carry out the analysis of the spectra, evaluation of the measurement data, whole body system type armchair with anthropomorphic phantom, whole body system of vertical scanning, distribution and location diagram of the internal dosimetry laboratory there are among the treated aspects. (Author

  16. Strengthening organizational performance through accreditation research-a framework for twelve interrelated studies: the ACCREDIT project study protocol

    Directory of Open Access Journals (Sweden)

    Pope Catherine

    2011-10-01

    Full Text Available Abstract Background Service accreditation is a structured process of recognising and promoting performance and adherence to standards. Typically, accreditation agencies either receive standards from an authorized body or develop new and upgrade existing standards through research and expert views. They then apply standards, criteria and performance indicators, testing their effects, and monitoring compliance with them. The accreditation process has been widely adopted. The international investments in accreditation are considerable. However, reliable evidence of its efficiency or effectiveness in achieving organizational improvements is sparse and the value of accreditation in cost-benefit terms has yet to be demonstrated. Although some evidence suggests that accreditation promotes the improvement and standardization of care, there have been calls to strengthen its research base. In response, the ACCREDIT (Accreditation Collaborative for the Conduct of Research, Evaluation and Designated Investigations through Teamwork project has been established to evaluate the effectiveness of Australian accreditation in achieving its goals. ACCREDIT is a partnership of key researchers, policymakers and agencies. Findings We present the framework for our studies in accreditation. Four specific aims of the ACCREDIT project, which will direct our findings, are to: (i evaluate current accreditation processes; (ii analyse the costs and benefits of accreditation; (iii improve future accreditation via evidence; and (iv develop and apply new standards of consumer involvement in accreditation. These will be addressed through 12 interrelated studies designed to examine specific issues identified as a high priority. Novel techniques, a mix of qualitative and quantitative methods, and randomized designs relevant for health-care research have been developed. These methods allow us to circumvent the fragmented and incommensurate findings that can be generated in small

  17. Neutron dosimetry in containment of a pressurized water reactor utilizing the Panasonic UD-802 dosimetry system

    International Nuclear Information System (INIS)

    The Panasonic UD-802 dosimeter was evaluated as a potential neutron dosimeter for use in containment of a PWR. The Panasonic UD-802 dosimeter, although designed as a beta and gamma dosimeter, is also sensitive to neutrons. UD-802 dosimeters were mounted on polyethylene phantoms and irradiated to known doses at selected locations in containment. The known neutron dose equivalents were determined based on remmeter dose rate measurements and stay times. The thermoluminescent response of the dosimeters and the known neutron dose equivalents were used to obtain a calibration factor at each location. The average calibration factor was 3.7 (unit of dosimeter response per mrem) and all calibration factors were within +-30% of this mean value. The dosimeter distance from the phantom was found to have minimal effect on the response but the system was directionally dependent, necessitating a correction in the calibration factor. The minimum significant dosimeter response was determined independent of any calibration factor. The minimum significant response of the UD-802 to neutrons is a function of the corresponding gamma exposure rate. It is concluded that the Panasonic UD-802 dosimeter can be used for neutron dosimetry in PWR containment

  18. Algorithm Verification for A TLD Personal Dosimetry System

    International Nuclear Information System (INIS)

    Dose algorithms are used in thermoluminescence personnel dosimetry for the interpretation of the dosimeter response in terms of equivalent dose. In the present study an Automated Harshaw 6600 reader was rigorously tested prior to use for dose calculation algorithm according to the standard established by the US Department of Energy Laboratory Accreditation Program (DOELAP). Also, manual Harshaw 4500 reader was used along with the ICRU slab phantom and the RANDO phantom in experimentally determining the photon personal doses in terms of deep dose, Hp(10), shallow dose, Hp(0.07), and eye lens dose, Hp(3),. Also, a Monte Carlo simulation program (VMC-dc) free code was used to simulate RANDO phantom irradiation process. The accuracy of the automated system lies well within DOELAP tolerance limits in all test categories

  19. 7 CFR 205.507 - Denial of accreditation.

    Science.gov (United States)

    2010-01-01

    ... Agriculture Regulations of the Department of Agriculture (Continued) AGRICULTURAL MARKETING SERVICE (Standards, Inspections, Marketing Practices), DEPARTMENT OF AGRICULTURE (CONTINUED) ORGANIC FOODS PRODUCTION ACT PROVISIONS NATIONAL ORGANIC PROGRAM Accreditation of Certifying Agents § 205.507 Denial of accreditation....

  20. NATIONAL ENVIRONMENTAL LABORATORY ACCREDITATION CONFERENCE (NELAC): CONSTITUTION, BYLAWS, AND STANDARDS

    Science.gov (United States)

    The principles and operating procedures for the National Environmental Laboratory Accreditation Conference (NELAC) are contained in the NELAC Constitution and Bylaws. The major portion of this document (standards) contains detailed requirements for accrediting environmental labo...

  1. Fast neutron dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    DeLuca, P.M. Jr.; Pearson, D.W.

    1992-01-01

    This progress report concentrates on two major areas of dosimetry research: measurement of fast neutron kerma factors for several elements for monochromatic and white spectrum neutron fields and determination of the response of thermoluminescent phosphors to various ultra-soft X-ray energies and beta-rays. Dr. Zhixin Zhou from the Shanghai Institute of Radiation Medicine, People's Republic of China brought with him special expertise in the fabrication and use of ultra-thin TLD materials. Such materials are not available in the USA. The rather unique properties of these materials were investigated during this grant period.

  2. INDIVIDUAL DOSIMETRY SERVICE

    CERN Multimedia

    2000-01-01

    Personnel in the distribution groups Aleph, Delphi, L3, Opal who also work for other experiments than at LEP, should contact their dispatchers to explain their activities for the future, after LEP dismantling in order to be maintained on the regular distribution list at Individual Dosimetry ServiceWe inform all staffs and users under regular dosimetric control that the dosimeters for the monitoring period JULY/AUGUST are available from their usual dispatchers.Please have your films changed before the 10th of July.The colour of the dosimeter valid in JULY/AUGUST is PINK.

  3. CNEA's (Comision Nacional de Energia Atomica) experience in the preparation of a national system for laboratory accreditation

    International Nuclear Information System (INIS)

    Within the regional markets, as it is the case of MERCOSUR , the laboratories which are suppliers of test and calibration results, are mutually recognized through the National Accreditation Systems. In Argentina there is a project to create a Center for the Accreditation of Test Laboratories. CNEA, which is involved in the execution of large projects and has adopted quality assurance criteria for a long time, requires for internal and external laboratories to be qualified. At the beginning of this year, a Committee for the Qualification of Laboratories was created in the Research and Development and Fuel Cycle Areas. Its objective was planning, management of documents, coordination, evaluation and quantification of laboratories, according to national IRAM and international ISO standards. This paper analyzes the organization of the system and the methods to evaluate and qualify laboratories as a process of growing up leading to the future National Accreditation System. (author). 3 figs

  4. Home infusion accreditation: another quality recognition branding tool.

    Science.gov (United States)

    Cannally, Sandra C

    2010-01-01

    Accreditation plays a significant role in fostering public confidence in the healthcare organization, enhancing organizational effectiveness, and improving patient care. It also provides the basis on which referral sources and payers can be assured that accredited organizations have complied with a common set of requirements and standards. Accreditation is the only comprehensive measure of a home infusion company's performance. The objective of this article is to provide information to help streamline and simplify the accreditation process and lower implementation times and associated costs.

  5. BUSINESS ETHICS AS AN ACCREDITATION REQUIREMENT: A KNOWLEDGE MAPPING APPROACH

    OpenAIRE

    Rita A. Franks; Albert D. Spalding, Jr

    2013-01-01

    Most of the more prominent and highly ranked business and management schools in the United States and elsewhere are accredited by one of two international accrediting organizations, the Association to Advance Collegiate Schools of Business (AACSB) or the Accreditation Council for Business Schools & Programs (ACBSP). Both of these organizations require the inclusion of business ethics in the curriculum of each accredited institution. Business ethics, however, is a concept that includes, overla...

  6. The importance of 3D dosimetry

    International Nuclear Information System (INIS)

    Radiation therapy has been getting progressively more complex for the past 20 years. Early radiation therapy techniques needed only basic dosimetry equipment; motorized water phantoms, ionization chambers, and basic radiographic film techniques. As intensity modulated radiation therapy and image guided therapy came into widespread practice, medical physicists were challenged with developing effective and efficient dose measurement techniques. The complex 3-dimensional (3D) nature of the dose distributions that were being delivered demanded the development of more quantitative and more thorough methods for dose measurement. The quality assurance vendors developed a wide array of multidetector arrays that have been enormously useful for measuring and characterizing dose distributions, and these have been made especially useful with the advent of 3D dose calculation systems based on the array measurements, as well as measurements made using film and portal imagers. Other vendors have been providing 3D calculations based on data from the linear accelerator or the record and verify system, providing thorough evaluation of the dose but lacking quality assurance (QA) of the dose delivery process, including machine calibration. The current state of 3D dosimetry is one of a state of flux. The vendors and professional associations are trying to determine the optimal balance between thorough QA, labor efficiency, and quantitation. This balance will take some time to reach, but a necessary component will be the 3D measurement and independent calculation of delivered radiation therapy dose distributions

  7. CNAL Successfully Passed APLAC Peer Evaluation Inspection Body Accreditation

    Institute of Scientific and Technical Information of China (English)

    2004-01-01

    @@ Aug. 9-12, 2004, APLAC (Asia Pacific Laboratory Accreditation Cooperation) conducted evaluation of CNAL on Inspection Body Accreditation. After four-day′s evaluation (including Secretariat Office and witnessing assessment), the evaluation group of APLAC declared that "CNAL has operated the Accreditation System of Inspection Body which complies with the requirements in MR001 and MR002.

  8. Improving Outcome Assessment in Information Technology Program Accreditation

    Science.gov (United States)

    Goda, Bryan S.; Reynolds, Charles

    2010-01-01

    As of March 2010, there were fourteen Information Technology programs accredited by the Accreditation Board for Engineering and Technology, known as ABET, Inc (ABET Inc. 2009). ABET Inc. is the only recognized institution for the accreditation of engineering, computing, and technology programs in the U.S. There are currently over 128 U.S. schools…

  9. Internal Dosimetry. Chapter 18

    International Nuclear Information System (INIS)

    The Committee on Medical Internal Radiation Dose (MIRD) is a committee within the Society of Nuclear Medicine. The MIRD Committee was formed in 1965 with the mission to standardize internal dosimetry calculations, improve the published emission data for radionuclides and enhance the data on pharmacokinetics for radiopharmaceuticals [18.1]. A unified approach to internal dosimetry was published by the MIRD Committee in 1968, MIRD Pamphlet No. 1 [18.2], which was updated several times thereafter. Currently, the most well known version is the MIRD Primer from 1991 [18.3]. The latest publication on the formalism was published in 2009 in MIRD Pamphlet No. 21 [18.4], which provides a notation meant to bridge the differences in the formalism used by the MIRD Committee and the International Commission on Radiological Protection (ICRP) [18.5]. The formalism presented in MIRD Pamphlet No. 21 [18.4] will be used here, although some references to the quantities and parameters used in the MIRD primer [18.3] will be made. All symbols, quantities and units are presented

  10. Calibrating a detector array for in vivo dosimetry beam electron In this work present a set of detectors calibration for measurements in vivo dose at the entrance of the radiation beam in electron beams; Calibracion de un conjunto de detectores para la dosimetria in vivo de haces de electrones

    Energy Technology Data Exchange (ETDEWEB)

    Martinez Jurado, T.; Jornet Sala, N.; Carrasco de Fez, P.; Eudaldo Puell, T.; Latorre Mussoll, A.; Rodriguez Latorre, D.; Ruiz Martinez, A.; Ribas Morales, M.

    2013-07-01

    In particular we have determined the variation in sensitivity of the diode depending on the field size, power, angle of incidence, DFS, temperature and cumulative dose. From these results we provide a methodology for calibrating a diode array for use in routine. The application of correction factors to reading diodes are warranted when the irradiation conditions differ from the reference by the level of tolerance among the expected dose and the dose obtained. (Author)

  11. Dosimetry on the radiological risks prevention in radiotherapy; La dosimetria en la prevencion de riesgos radiologicos en radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Fornet R, O. M.; Perez G, F., E-mail: nuclear2@citmahlg.holguin.inf.cu [Delegacion Territorial del CITMA, Peralta 16 esq. P. Feria, Rpto. Peralta, 80400 Holguin (Cuba)

    2014-08-15

    Dosimetry in its various forms plays a determining role on the radiological risks prevention in radiotherapy. To prove this in this paper is shown an analysis based on the risk matrix method, how the dosimetry can influence in each stages of a radiotherapy service; installation and acceptance, operation, maintenance and calibration. For each one of these stages the role that can play is analyzed as either the initiating event of a radiological accident or limiting barrier of these events of the dosimetric processes used for the individual dosimetry, the area monitoring, fixed or portable, for radiation beam dosimetry and of the patients for a radiotherapy service with cobalt-therapy equipment. The result of the study shows that the application of a prospective approach in the role evaluation of dosimetry in the prevention and mitigation of the consequences of a radiological accident in radiotherapy is crucial and should be subject to permanent evaluation at each development stage of these services. (author)

  12. ESR/alanine dosimetry applied to radiation processing

    International Nuclear Information System (INIS)

    The radiation processing of food products is specified in terms of absorbed dose, and processing quality is assessed on the basis of absorbed dose measurements. The validity of process quality control is highly dependent on the quality of the measurements and associated instrumentation; in this respect, dosimetry calibration by an Organization with official status provides an essential guarantee of validity to the quality control steps taken. The Laboratoire de Metrologie des Rayonnements Ionisants (L.M.R.I.) is the primary standards and evaluation laboratory approved by the Bureau National de Metrologie (B.N.M.), which is the French National Bureau of Standards. The LMRI implements correlation procedures in response to the various requirements which arise in connection with high doses and doserates. Such procedures are mainly based on ESR/alanine spectrometry, a dosimetry technique ideally suited to that purpose. Dosemeter geometry and design are tailored to operating conditions. Photon dosemeters consist of a detector material in powder or compacted form, and a wall with thickness and chemical composition consistent with the application. Electron dosemeters have a detector core of compacted alanine with thickness down to a few tenths of a millimeter. The ESR/alanine dosimetry technique, developed at LMRI is a flexible, reliable and accurate tool which effectively meets the various requirements arising in the field of reference dosimetry, where high doses and doserates are involved. (author)

  13. External quality audit programmes for radiotherapy dosimetry and equipment

    International Nuclear Information System (INIS)

    It is widely accepted that individual radiotherapy centres should have in place a comprehensive quality assurance programme on all the necessary steps for the delivery of safe accurate treatment. As regards the performance of radiotherapy equipment and dosimetry, the most widely used process of external checking has been dosimetry intercomparison, comparing independently measured doses to locally stated doses in a variety of conditions. These have been at a number of different levels: from basic beam calibration; up to and including exercises employing anatomic or pseudo-anatomic phantoms and incorporating tests of treatment planning equipment and procedures. Some of these have been one-off exercises, whilst others are continuing, or have given rise to on-going quality audit programmes on a national (or wider) basis. A number of these have evolved, or are evolving, into audits which include external checking of the achievement of standards in performance of treatment equipment, as well as in the dosimetry in each institution involved. The principles and methodologies of the various types of external checking programmes for treatment equipment and dosimetry are reviewed, covering the experimental approaches and the tolerances applied. What is included in a given programme will, of necessity, depend on the resources available and the purpose of the exercise. Methods and tolerances must be matched to endpoint. Tolerance levels must take into account the experimental uncertainties of the measurement methods employed. Finally, external audit can only be used to complement, and in conjunction with, institutional quality assurance programmes and not as a substitute for them

  14. Approaches to the ISO/IEC 17025 accreditation for Pu and U accountancy analysis

    International Nuclear Information System (INIS)

    The quality control section of Plutonium Fuel Development Center (PFDC) of Japan Atomic Energy Agency has been analyzing isotopic compositions by Mass Spectrometry as well as content by Isotope Dilution Mass Spectrometry (IDMS) of plutonium and uranium in nuclear materials. Along with establishing and managing the quality assurance system, ensuring the reliability of the analysis data is important. PFDC has been establishing the quality management system with ISO9001. ISO9001 consists of management requirements for quality system of organizations. While ISO/IEC 17025 consists of technical requirements for the competence of testing and calibration laboratories in addition to the management requirements. The quality control section addressed technical improvement to improve further reliability of analysis quality and we have accredited for ISO/IEC 17025 of isotopic compositions and content of plutonium and uranium in nuclear materials in March 2010. In this presentation, we report our approaches to the ISO/IEC 17025 accreditation and operation status. (author)

  15. Information from the Dosimetry Service

    CERN Multimedia

    2006-01-01

    Please note the following opening hours of the Service: From 31st July onwards: Every morning from 8:30 to 12:00 The Service is closed in the afternoons. We should like to remind you that dosimeters cannot be sent to customers by internal mail. Short-term dosimeters (VCTs) must always be returned to the Service after use and must not be left on the racks in the experimental areas or in the secretariats. Dosimetry Service Tel 72155 Bldg. 24 E 011 Dosimetry.service@cern.ch http://cern.ch/rp-dosimetry

  16. Comparison between IAEA/TRS-277 and IAEA/TRS-398 protocols for electron beam dosimetry with cylindrical ionization chambers

    International Nuclear Information System (INIS)

    With the purpose to guarantee an uncertainty in the dosimetry in radiation therapy, the International Atomic Energy Agency (IAEA) published in 1987 the Technical Reports Series (TRS) number 277 - Absorbed Dose Determination in Photon and Electron Beams - An International Code of Practice -, updated in 1997, when was published its second edition. In 2000 was published the TRS number 398 - Absorbed Dose Determination in External Beam Radiotherapy - An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. The TRS number 398 brings a great conceptual change in relation to the basis of the formalism, before based on calibration factor in terms of air kerma, and now based on calibration factor in terms of absorbed dose in water. Since the TRS number 398 was published, the Secondary Standard Dosimetry Laboratories are calibrating the user's ionization chambers in terms of absorbed dose to water. However, nor all the clinics in Rio de Janeiro and Brazil have its ionization chambers calibrated in terms of absorbed dose to water. The National Cancer Institute, where the measurements were taken, was the first institution in the Rio de Janeiro to have its ionization chambers calibrated in terms of a new formalism. This work describes a comparison between dosimetry done with a cylindrical ionization chamber under electron beams utilizing the TRS number 277 formalism, based on air kerma, and the TRS number 398 formalism, based on absorbed dose to water, reporting the uncertainties variation of the dosimetry associated to each protocol. (author)

  17. 38 CFR 21.4253 - Accredited courses.

    Science.gov (United States)

    2010-07-01

    ... teacher's certificate or teacher's degree. (5) The course is approved by the State as meeting the... college degree) or it may be vocational or professional (an occupation). (c) Accrediting agencies. A... student is to graduate. For example, a 4-year college may require a 1.5 grade point average the first...

  18. Assessment and accreditation of Mexican medical schools.

    Science.gov (United States)

    Cordova, J A; Aguirre, E; Hernández, A; Hidalgo, V; Domínguez, F; Durante, I; Jesús, R; Castillo, O

    1996-09-01

    With the objective of evaluating and accrediting the quality of medical education in the country, the Mexican Association of Medical Schools initiated the National Programme for the Strengthening of the Quality of Medical Education (PNFCE). This programme led to the establishment of the National System of Accreditation. Medical school deans in Mexico determined the criteria for the evaluation of quality and its subsequent standards through a consensus process. The following 10 criteria resulted: general basis and educational objectives; government and institutional orientation; educational programme and academic structure; educational process assessment; students; teaching staff; institutional coherence; resources; clinical sites; and administration. Eighty-eight standards were developed in the instrument designed for the self-evaluation phase. The information resulting from the self-evaluation will be verified by a group of experts during a survey visit, which will be finalized with a report to serve as the basis for the decision to be made by the Accreditation Commission. The self-evaluation phase started in 1994. In 1996 four schools submitted their request for accreditation. As of July 1996, one survey visit has been completed and three more are programmed for the second half of the year.

  19. 42 CFR 422.157 - Accreditation organizations.

    Science.gov (United States)

    2010-10-01

    ....157 Public Health CENTERS FOR MEDICARE & MEDICAID SERVICES, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) MEDICARE PROGRAM MEDICARE ADVANTAGE PROGRAM Quality Improvement § 422.157 Accreditation..., it applies and enforces standards that are at least as stringent as Medicare requirements...

  20. Accreditation's Benefits for Individuals and Institutions

    Science.gov (United States)

    McGuire, Patricia A.

    2009-01-01

    Participation in accreditation processes, on visiting teams as well as through institutional self-study, is an excellent opportunity for individual academics to augment their professional expertise in a range of higher education issues: strategic planning and assessment, resource management and capital investments, curriculum planning and program…

  1. Does accreditation stimulate change? A study of the impact of the accreditation process on Canadian healthcare organizations

    Directory of Open Access Journals (Sweden)

    Shabah Abdo

    2010-04-01

    Full Text Available Abstract Background One way to improve quality and safety in healthcare organizations (HCOs is through accreditation. Accreditation is a rigorous external evaluation process that comprises self-assessment against a given set of standards, an on-site survey followed by a report with or without recommendations, and the award or refusal of accreditation status. This study evaluates how the accreditation process helps introduce organizational changes that enhance the quality and safety of care. Methods We used an embedded multiple case study design to explore organizational characteristics and identify changes linked to the accreditation process. We employed a theoretical framework to analyze various elements and for each case, we interviewed top managers, conducted focus groups with staff directly involved in the accreditation process, and analyzed self-assessment reports, accreditation reports and other case-related documents. Results The context in which accreditation took place, including the organizational context, influenced the type of change dynamics that occurred in HCOs. Furthermore, while accreditation itself was not necessarily the element that initiated change, the accreditation process was a highly effective tool for (i accelerating integration and stimulating a spirit of cooperation in newly merged HCOs; (ii helping to introduce continuous quality improvement programs to newly accredited or not-yet-accredited organizations; (iii creating new leadership for quality improvement initiatives; (iv increasing social capital by giving staff the opportunity to develop relationships; and (v fostering links between HCOs and other stakeholders. The study also found that HCOs' motivation to introduce accreditation-related changes dwindled over time. Conclusions We conclude that the accreditation process is an effective leitmotiv for the introduction of change but is nonetheless subject to a learning cycle and a learning curve. Institutions invest

  2. Current status of internal dosimetry service laboratories in Brazil

    Energy Technology Data Exchange (ETDEWEB)

    Dantas, B.M.; Acar, M.E.D.; Dantas, A.L.A. [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil). In Vivo Monitoring Lab.], e-mail: bmdantas@ird.gov.br; Cardoso, J.C.S. [Instituto de Pesquisas Energeticas e Nucleares (IPEN-CNEN/SP), Sao Paulo, SP (Brazil). In Vivo Monitoring Lab.], e-mail: jcardoso@ipen.br; Tadei, M.H.T. [Laboratorio de Pocos de Caldas, (LAPOC/CNEN-MG), MG (Brazil). Radiochemistry Lab.], e-mail: mhtaddei@cnen.gov.br; Juliao, L.Q.C. [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil). In Vitro Bioassay Lab.], e-mail: ligia@ird.gov.br; Lima, M.F. [Instituto de Pesquisas Energeticas e Nucleares (IPEN-CNEN/SP), Sao Paulo, SP (Brazil). Radiochemistry Lab.], e-mail: mflima@ipen.br; Arine, D.R. [Centro Tecnologico da Marinha em Sao Paulo (CTMSP) , SP (Brazil). Radioecology Lab.], e-mail: djair@ctmsp.mar.mil.br; Ramos, M.A. [Central Nuclear Almirante Alvaro Alberto (CNAAA), Angra dos Reis, RJ (Brazil). In Vivo Monitoring Lab.], e-mail: mramos@eletronuclear.gov.br; Alonso, T.C. [Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil). In Vivo Monitoring Lab.], e-mail: alonso@cdtn.br; Lima, F.F. [Centro Regional de Ciencias Nucleares (CRCN-NE/CNEN-PE), Recife, PE (Brazil). In Vivo Monitoring Lab.], e-mail: fflima@cnen.gov.br

    2009-07-01

    Several technical documents related to internal dosimetry have been released by the IAEA and ICRP after nuclear and radiological accidents, such as the Chernobyl and Goiania. However, standard bioassay procedures and methodologies for bioassay data interpretation are still under discussion and, in some cases, not well understood by the professionals involved in this specific field of radiation protection. Therefore, both in routine and emergency monitoring, responses may differ markedly among Dosimetry Laboratories and it may be difficult to interpret and use the bioassay data generated. The resulting misunderstanding can impair countermeasures and remediation operations and enhance significant socio-economic and political consequences. Currently it is recognized worldwide the need to have a realistic evaluation of the reliability of the services provided by specific laboratory as well as a clear compliance with best practices and a permanent effort to improve data interpretation. The objective of this work is to ensure regular and systematic quality monitoring of the Accredited Laboratory Network composed by the Brazilian governmental Institutes which will comprise expert teams able to provide, upon request, reliable services in case of a radiological accidents and follow-up operations, as well as internal dose evaluation of occupationally exposed workers. (author)

  3. Calibration uncertainty

    DEFF Research Database (Denmark)

    Heydorn, Kaj; Anglov, Thomas

    2002-01-01

    uncertainty was verified from independent measurements of the same sample by demonstrating statistical control of analytical results and the absence of bias. The proposed method takes into account uncertainties of the measurement, as well as of the amount of calibrant. It is applicable to all types......Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration...

  4. Result of China National Accreditation for Laboratories Achieving Good Credit

    Institute of Scientific and Technical Information of China (English)

    2005-01-01

    @@ With the efforts of the past 10 years, the accreditation system for laboratories, which is complete in range, integrate on function and normative on operation, has been set up in China, and has achieved good reputation home and abroad. The result of laboratories accreditation is widely admitted and used in the international. Since the China national accreditation system for laboratories entering the international multilateral mutual recognition arrangement, the accreditation for laboratories has been playing an increasingly important role in the international trade. The testing result provided by the accreditation laboratories is required by many international-famous enterprises when they purchase in China, and there have been 37 economic systems admitting the result of China national accreditation for laboratories. More and more governmental departments require using accreditation for laboratories in the administrative management and law enforcement.

  5. Internal dosimetry technical basis manual

    International Nuclear Information System (INIS)

    The internal dosimetry program at the Savannah River Site (SRS) consists of radiation protection programs and activities used to detect and evaluate intakes of radioactive material by radiation workers. Examples of such programs are: air monitoring; surface contamination monitoring; personal contamination surveys; radiobioassay; and dose assessment. The objectives of the internal dosimetry program are to demonstrate that the workplace is under control and that workers are not being exposed to radioactive material, and to detect and assess inadvertent intakes in the workplace. The Savannah River Site Internal Dosimetry Technical Basis Manual (TBM) is intended to provide a technical and philosophical discussion of the radiobioassay and dose assessment aspects of the internal dosimetry program. Detailed information on air, surface, and personal contamination surveillance programs is not given in this manual except for how these programs interface with routine and special bioassay programs

  6. Radioembolization Dosimetry: The Road Ahead

    Energy Technology Data Exchange (ETDEWEB)

    Smits, Maarten L. J., E-mail: m.l.j.smits-3@umcutrecht.nl; Elschot, Mattijs [University Medical Center Utrecht, Department of Radiology and Nuclear Medicine (Netherlands); Sze, Daniel Y. [Stanford University School of Medicine, Division of Interventional Radiology (United States); Kao, Yung H. [Austin Hospital, Department of Nuclear Medicine (Australia); Nijsen, Johannes F. W. [University Medical Center Utrecht, Department of Radiology and Nuclear Medicine (Netherlands); Iagaru, Andre H. [Stanford University School of Medicine, Division of Nuclear Medicine and Molecular Imaging (United States); Jong, Hugo W. A. M. de; Bosch, Maurice A. A. J. van den; Lam, Marnix G. E. H. [University Medical Center Utrecht, Department of Radiology and Nuclear Medicine (Netherlands)

    2015-04-15

    Methods for calculating the activity to be administered during yttrium-90 radioembolization (RE) are largely based on empirical toxicity and efficacy analyses, rather than dosimetry. At the same time, it is recognized that treatment planning based on proper dosimetry is of vital importance for the optimization of the results of RE. The heterogeneous and often clustered intrahepatic biodistribution of millions of point-source radioactive particles poses a challenge for dosimetry. Several studies found a relationship between absorbed doses and treatment outcome, with regard to both toxicity and efficacy. This should ultimately lead to improved patient selection and individualized treatment planning. New calculation methods and imaging techniques and a new generation of microspheres for image-guided RE will all contribute to these improvements. The aim of this review is to give insight into the latest and most important developments in RE dosimetry and to suggest future directions on patient selection, individualized treatment planning, and study designs.

  7. Internal dosimetry technical basis manual

    Energy Technology Data Exchange (ETDEWEB)

    1990-12-20

    The internal dosimetry program at the Savannah River Site (SRS) consists of radiation protection programs and activities used to detect and evaluate intakes of radioactive material by radiation workers. Examples of such programs are: air monitoring; surface contamination monitoring; personal contamination surveys; radiobioassay; and dose assessment. The objectives of the internal dosimetry program are to demonstrate that the workplace is under control and that workers are not being exposed to radioactive material, and to detect and assess inadvertent intakes in the workplace. The Savannah River Site Internal Dosimetry Technical Basis Manual (TBM) is intended to provide a technical and philosophical discussion of the radiobioassay and dose assessment aspects of the internal dosimetry program. Detailed information on air, surface, and personal contamination surveillance programs is not given in this manual except for how these programs interface with routine and special bioassay programs.

  8. Medical dosimetry in Hungary

    Science.gov (United States)

    Turák, O.; Osvay, M.; Ballay, L.

    2012-09-01

    Radiation exposure of medical staff during cardiological and radiological procedures was investigated. The exposure of medical staff is directly connected to patient exposure. The aim of this study was to determine the distribution of doses on uncovered part of body of medical staff using LiF thermoluminescent (TL) dosimeters in seven locations. Individual Kodak film dosimeters (as authorized dosimetry system) were used for the assessment of medical staff's effective dose. Results achieved on dose distribution measurements confirm that wearing only one film badge under the lead apron does not provide enough information on the personal dose. The value of estimated annual doses on eye lens and extremities (fingers) were in good correlation with international publications.

  9. Dosimetry of iodoantipyrine

    International Nuclear Information System (INIS)

    Dosimetry of iodoantipyrine labeled with radioactive iodine was determined by measuring the biodistribution of 131I-iodoantipyrine in 41 female rabbits. Following administration of the radiopharmaceutical, subjects were killed at 0.5, 6, 12, 17, 24, 36, and 48 h. Organs and samples of tissues and body fluids were assayed. Results were corrected for physical decay. Exponential functions were employed to describe the time-concentration curves; representative value would be the biological half life of 9.96±0.55 h for blood. Cumulated activity estimates for 123I, 125I and 131I were then computed. Extrapolation to absorbed dose in humans followed the formulation of the Medical International Radiation Dose (MIRD) Committee of the Society of Nuclear Medicine. The whole body absorbed doses are 0.7 μGray, 0.5 μGray and 2.9 μGray per MBq of 123I, 123I, and 131I administered respectively. (orig.)

  10. Strahlungsmessung und Dosimetrie

    CERN Document Server

    Krieger, Hanno

    2013-01-01

    „Strahlungsquellen und Dosimetrie“ ist Teil einer Lehrbuchreihe zur Strahlungsphysik und zum Strahlenschutz. Der erste Teil befasst sich mit den physikalischen Grundlagen der Strahlungsdetektoren und der Strahlungsmessung. Im zweiten Teil werden die Konzepte und Verfahren der klinischen Dosimetrie dargestellt. Der dritte Abschnitt erläutert ausführlich die Dosisverteilungen der klinisch angewendeten Strahlungsarten. Im vierten Teil werden weitere Messaufgaben der Strahlungsphysik einschließlich der Messsysteme für die Bildgebung mit Röntgenstrahlung dargestellt. Neben den grundlegenden Ausführungen enthält dieser Band im laufenden Text zahlreiche Tabellen und Grafiken zur technischen und medizinischen Radiologie, die bei der praktischen Arbeit sehr hilfreich sein können und 199 Übungsaufgaben mit Lösungen zur Vertiefung der Inhalte. Für die zweite Auflage wurden die Darstellungen der Elektronen- und der Protonendosimetrie sowie der bildgebenden Verfahren mit Computertomografen deutlich erweit...

  11. Hanford internal dosimetry program manual

    International Nuclear Information System (INIS)

    This document describes the Hanford Internal Dosimetry program. Program Services include administrating the bioassay monitoring program, evaluating and documenting assessments of internal exposure and dose, ensuring that analytical laboratories conform to requirements, selecting and applying appropriate models and procedures for evaluating internal radionuclide deposition and the resulting dose, and technically guiding and supporting Hanford contractors in matters regarding internal dosimetry. 13 refs., 16 figs., 42 tabs

  12. Fifth international radiopharmaceutical dosimetry symposium

    International Nuclear Information System (INIS)

    This meeting was held to exchange information on how to get better estimates of the radiation absorbed dose. There seems to be a high interest of late in patient dosimetry; discussions were held in the light of revised risk estimates for radiation. Topics included: Strategies of Dose Assessment; Dose Estimation for Radioimmunotherapy; Dose Calculation Techniques and Models; Dose Estimation for Positron Emission Tomography (PET); Kinetics for Dose Estimation; and Small Scale Dosimetry and Microdosimetry. (VC)

  13. Hanford internal dosimetry program manual

    Energy Technology Data Exchange (ETDEWEB)

    Carbaugh, E.H.; Sula, M.J.; Bihl, D.E.; Aldridge, T.L.

    1989-10-01

    This document describes the Hanford Internal Dosimetry program. Program Services include administrating the bioassay monitoring program, evaluating and documenting assessments of internal exposure and dose, ensuring that analytical laboratories conform to requirements, selecting and applying appropriate models and procedures for evaluating internal radionuclide deposition and the resulting dose, and technically guiding and supporting Hanford contractors in matters regarding internal dosimetry. 13 refs., 16 figs., 42 tabs.

  14. Dosimetry in radionuclide therapy

    International Nuclear Information System (INIS)

    While it is known that therapeutic effects of radionuclides are due to absorbed radiation dose and to radiosensitivity, individual dosimetry in 'Gy' is practiced rarely in clinical Nuclear Medicine but 'doses' are described in 'mCi' or 'MBq', which is only indirectly related to 'Gy' in the target. To estimate 'Gy', the volume of the target, maximum concentration of the radiopharmaceutical in it and residence time should be assessed individually. These parameters can be obtained usually only with difficulty, involving possibly also quantitative SPET or PET, modern imaging techniques (sonography, CT, MRT), substitution of y- or positron emitting radiotracers for β-emitting radiopharmaceuticals as well as whole-body distribution studies. Residence time can be estimated by obtaining data on biological half-life of a comparable tracer and transfer of these data in the physical characteristics of the therapeutic agent. With all these possibilities for gross dosimetry the establishment of a dose-response-relation should be possible. As distribution of the radiopharmaceutical in lesions is frequently inhomogenous and microdosimetric conditions are difficult to assess in vivo as yet, it could be observed since decades that empirically set, sometimes 'fixed' doses (mCi or MBq) can also be successful in many diseases. Detailed dosimetric studies, however, are work- and cost-intensive. Nevertheless, one should be aware at a time when more sophisticated therapeutic possibilities in Nuclear Medicine arise, that we should try to estimate radiation dose (Gy) in our new methods even as differences in individual radiosensitivity cannot be assessed yet and studies to define individual radiosensitivity in lesions should be encouraged. (author)

  15. Dosimetry in intravascular brachytherapy

    International Nuclear Information System (INIS)

    Among the cardiovascular diseases responsible for deaths in the adult population in almost all countries of the world, the most common is acute myocardial infarction, which generally occurs because of the occlusion of one or more coronary arteries. Several diagnostic techniques and therapies are being tested for the treatment of coronary artery disease. Balloon angioplasty has been a popular treatment which is less invasive than traditional surgeries involving revascularization of the myocardium, thus promising a better quality of life for patients. Unfortunately, the rate of restenosis (re-closing of the vessel) after balloon angioplasty is high (approximately 30-50% within the first year after treatment).Recently, the idea of delivering high radiation doses to coronary arteries to avoid or delay restenosis has been suggested. Known as intravascular brachytherapy, the technique has been used with several radiation sources, and researchers have obtained success in decreasing the rate of restenosis in some patient populations. In order to study the radiation dosimetry in the patient and radiological protection for the attending staff for this therapy, radiation dose distributions for monoenergetic electrons and photons (at nine discrete energies) were calculated for blood vessels of diameter 0.15, o,30 and 0.45 cm with balloon and wire sources using the radiation transport code MCNP4B. Specific calculations were carried out for several candidate radionuclides as well. Two s tent sources (metallic prosthesis that put inside of patient's artery through angioplasty) employing 32 P are also simulated. Advantages and disadvantages of the various radionuclides and source geometries are discussed. The dosimetry developed here will aid in the realization of the benefits obtained in patients for this promising new technology. (author)

  16. The Vinca dosimetry experiment

    International Nuclear Information System (INIS)

    On 15 October 1958 there occurred a very brief uncontrolled run of the zero-power reactor at the Boris Kidric Institute of Nuclear Science, Vinca, near Belgrade, Yugoslavia. During this run six persons received various doses of radiation. They were subsequently given medical treatment of a novel kind at the Curie Hospital, Paris. In atomic energy operations to date, very few accidents involving excessive radiation exposure to human beings have occurred. In fact, the cases of acute radiation injury are limited to about 30 known high exposures, few of which were in the lethal or near-lethal range. Since direct experiment to determine the effects of ionizing radiation on man is unacceptable, information on these effects has to be based on a consideration of data relating to accidental exposures, viewed in the light of the much more extensive data obtained from experiments on animals. Therefore, any direct information on the effects of radiation on humans is very valuable. The international dosimetry project described in this report was carried out at Vinca, Yugoslavia, under the auspices of the International Atomic Energy Agency to determine the precise amount of radiation to which the persons had been exposed during the accident. These dosimetry data, together with the record of the carefully observed clinical effects, are of importance both for the scientific study of radiation effects on man and for the development of methods of therapy. The experiment and measurements were carried out at the end of April 1960. The project formed part of the Agency's research programme in the field of health and safety. The results of the experiment are made available through this report to all Member States

  17. Internal dosimetry, past and future

    International Nuclear Information System (INIS)

    This paper is a review of the progress in the dosimetry of internally deposited radionuclides (internal dosimetry) since World War II. Previous to that, only naturally occurring radionuclides were available and only a limited number of studies of biokinetics and dosimetry were done. The main radionuclides studied were 226Ra, 228Ra, and 224Ra but natural uranium was also studied mainly because of its toxic effect as a heavy metal, and not because it was radioactive. The effects of 226Ra in bone, mainly from the radium dial painters, also formed the only bases for the radiotoxicity of radionuclides in bone for many years, and it is still, along with 224Ra, the main source of information on the effects of alpha emitters in bone. The publications of the International Commission on Radiological Protection that have an impact on internal dosimetry are used as mileposts for this review. These series of publications, more than any other, represent a broad consensus of opinion within the radiation protection community at the time of their publication, and have formed the bases for radiation protection practice throughout the world. This review is not meant to be exhaustive; it is meant to be a personnel view of the evolution of internal dosimetry, and to present the author's opinion of what the future directions in internal dosimetry will be. 39 refs., 2 tabs

  18. On multichannel film dosimetry with channel-independent perturbations

    CERN Document Server

    Méndez, Ignasi; Hudej, Rihard; Strojnik, Andrej; Casar, Božidar

    2014-01-01

    Purpose: Different multichannel methods for film dosimetry have been proposed in the literature. Two of them are the weighted mean method and the method put forth by Micke et al and Mayer et al. The purpose of this work was to compare their results and to develop a generalized channel-independent perturbations framework in which both methods enter as special cases. Methods: Four models of channel-independent perturbations were compared: weighted mean, Micke-Mayer method, uniform distribution and truncated normal distribution. A closed-form formula to calculate film doses and the associated Type B uncertainty for all four models was deduced. To evaluate the models, film dose distributions were compared with planned and measured dose distributions. At the same time, several elements of the dosimetry process were compared: film type EBT2 versus EBT3, different waiting-time windows, reflection mode versus transmission mode scanning, and planned versus measured dose distribution for film calibration and for gamma-...

  19. Alternative statistical methods for cytogenetic radiation biological dosimetry

    CERN Document Server

    Fornalski, Krzysztof Wojciech

    2014-01-01

    The paper presents alternative statistical methods for biological dosimetry, such as the Bayesian and Monte Carlo method. The classical Gaussian and robust Bayesian fit algorithms for the linear, linear-quadratic as well as saturated and critical calibration curves are described. The Bayesian model selection algorithm for those curves is also presented. In addition, five methods of dose estimation for a mixed neutron and gamma irradiation field were described: two classical methods, two Bayesian methods and one Monte Carlo method. Bayesian methods were also enhanced and generalized for situations with many types of mixed radiation. All algorithms were presented in easy-to-use form, which can be applied to any computational programming language. The presented algorithm is universal, although it was originally dedicated to cytogenetic biological dosimetry of victims of a nuclear reactor accident.

  20. Computer simulations for internal dosimetry using voxel models.

    Science.gov (United States)

    Kinase, Sakae; Mohammadi, Akram; Takahashi, Masa; Saito, Kimiaki; Zankl, Maria; Kramer, Richard

    2011-07-01

    In the Japan Atomic Energy Agency, several studies have been conducted on the use of voxel models for internal dosimetry. Absorbed fractions (AFs) and S values have been evaluated for preclinical assessments of radiopharmaceuticals using human voxel models and a mouse voxel model. Computational calibration of in vivo measurement system has been also made using Japanese and Caucasian voxel models. In addition, for radiation protection of the environment, AFs have been evaluated using a frog voxel model. Each study was performed by using Monte Carlo simulations. Consequently, it was concluded that these data of Monte Carlo simulations and voxel models could adequately reproduce measurement results. Voxel models were found to be a significant tool for internal dosimetry since the models are anatomically realistic. This fact indicates that several studies on correction of the in vivo measurement efficiency for the variability of human subjects and interspecies scaling of organ doses will succeed.

  1. Effects of dose fractionation on the response of alanine dosimetry

    International Nuclear Information System (INIS)

    Alanine dosimetry is well established as a transfer standard and is becoming more prevalently used in routine dosimetry systems for radiation processing. Many routine measurement applications in radiation processing involve absorbed dose measurements resulting from fractioned exposures to ionizing radiation. Fractioning of absorbed dose is identified as an influence quantity (ISO/ASTM, 2013). This paper reports on study results of absorbed dose fractioning characteristics of alanine for gamma and high energy electron beam radiation sources. The results of this study indicate a radiation response difference due to absorbed dose fractioning in response can be observed after four fractionations for high-energy electron beams and no difference up to seven fractions for gamma rays using an ANOVA evaluation method. - Highlights: • Fractioning effects signaled in electron beam using an ANOVA at 6 equal increments. • Fractioning effects not signaled in gamma using an ANOVA up to 7 equal increments. • Insensitivity of alanine to dose fractioning indicates nominal impact on calibration

  2. On the operational quantity Hp(3) for eye lens dosimetry

    International Nuclear Information System (INIS)

    In the past the operational quantity Hp(3) has been defined for calibration purposes in a slab phantom. Recently, an additional phantom in the form of a cylinder was suggested for eye lens dosimetry as a cylinder much better approximates the shape of a head than a slab. Therefore, this work investigates whether the quantity Hp(3), when defined in the respective calibration phantom, adequately estimates the eye lens dose (or is at least conservative) depending on the phantom: it turns out that in most cases both calibration phantoms are similarly well suited. Finally, the definition of the eye lens dose is discussed together with possible consequences on the definition of Hp(3): the consideration of only the radiation sensitive volume of the lens causes Hp(3) not to be conservative in beta radiation fields. (paper)

  3. Hanford External Dosimetry Technical Basis Manual PNL-MA-842

    Energy Technology Data Exchange (ETDEWEB)

    Rathbone, Bruce A.

    2010-01-01

    manual by PNNL was discontinued beginning with Revision 0.2. Revision Log: Rev. 0 (2/25/2005) Major revision and expansion. Rev. 0.1 (3/12/2007) Updated Chapters 5, 6 and 9 to reflect change in default ring calibration factor used in HEDP dose calculation software. Factor changed from 1.5 to 2.0 beginning January 1, 2007. Pages on which changes were made are as follows: 5.23, 5.69, 5.78, 5.80, 5.82, 6.3, 6.5, 6.29, and 9.2. Rev 0.2 (8/28/2009) Updated Chapters 3, 5, 6, 8 and 9. Chapters 6 and 8 were significantly expanded. References in the Preface and Chapters 1, 2, 4, and 7 were updated to reflect updates to DOE documents. Approved by HPDAC on 6/2/2009. Rev 1.0 (1/1/2010) Major revision. Updated all chapters to reflect the Hanford site wide implementation on January 1, 2010 of new DOE requirements for occupational radiation protection. The new requirements are given in the June 8, 2007 amendment to 10 CFR 835 Occupational Radiation Protection (Federal Register, June 8, 2007. Title 10 Part 835. U.S., Code of Federal Regulations, Vol. 72, No. 110, 31904-31941). Revision 1.0 to the manual replaces ICRP 26 dosimetry concepts and terminology with ICRP 60 dosimetry concepts and terminology and replaces external dose conversion factors from ICRP 51 with those from ICRP 74 for use in measurement of operational quantities with dosimeters. Descriptions of dose algorithms and dosimeter response characteristics, and field performance were updated to reflect changes in the neutron quality factors used in the measurement of operational quantities.

  4. New instrument calibration facility for the DOE Savannah River Site

    Energy Technology Data Exchange (ETDEWEB)

    Wilkie, W.H.; Polz, E.J. [Westinghouse Savannah River Company, Aiken, SC (United States)

    1993-12-31

    A new laboratory facility is being designed, constructed, and equipped at the Savannah River Site (SRS) as a fiscal year 1992 line item project. This facility will provide space and equipment for test, evaluation, repair, maintenance, and calibration of radiation monitoring instrumentation. The project will replace an obsolete facility and will allow implementation of program upgrades necessary to meet ANSI N323 requirements and National Voluntary Laboratory Accreditation Program (NVLAP) criteria for accreditation of federally owned secondary calibration laboratories. An outline of the project is presented including description, scope, cost, management organization, chronology, and current status. Selected design criteria and their impacts on the project are discussed. The upgraded SRS calibration program is described, and important features of the new facility and equipment that will accommodate this program are listed. The floor plan for the facility is shown, and equipment summaries and functional descriptions for each area are provided.

  5. Accreditation in university environmental radioactivity laboratories

    International Nuclear Information System (INIS)

    Full text: The experimental work performed in university laboratories comes from many different fields and it is assumed to be of high quality. In general, the results are published in national or international journals or presented at conferences. Only a few laboratories have a clear understanding of the importance of implementing Quality Assurance Systems and the accreditation of their activities according to the international standards, such as ISO 17025. Today, few universities include this issue in the ir programmes. Most laboratories associate quality assurance with the fact that referees before publication have revised their works. Here the authors describe their experience in two university laboratories involved in environmental radioactivity control. Both laboratories have implanted a Quality Assurance System based on ISO 17025, the standard used for accreditation of the technical competence of laboratories. One of them (LARA-UPC) belongs to a research institute and the other (LRAUB) belongs to a university department with different logistic organisation. Both laboratories provide services to public and private institutions along side their teaching and research activities. The Quality Assurance Unit (UGQ-UB) is responsible for activities related to technical support in implementation and assessment in quality systems. In the case of these laboratories this UGQ performs internal audits. Accreditation is particularly important in environmental radioactivity analysis, where objective evidence of the quality of the data is required. Moreover, the results of radioactivity analysis are important: e.g. quality of water for human consumption (Directive 98/83/CE), environmental surveys (PVRA, Art. 35 of the Euratom Treaty for EU members), imports of agricultural products (Directive 99/1661/EC), export certificate required for agricultural products (2001/1621/EC), measurements in support of health and safety. It is important to assure the accuracy and precision of

  6. [Merits of acquiring ISO15189 accreditation].

    Science.gov (United States)

    Kitagawa, Masami

    2010-01-01

    In Japan, an ISO15189 accreditation system was started in 2005. To date, 47 hospitals have been accredited. In this session, I will present the merits of acquiring accreditation regarding ISO15189 based on our experience. Our hospital has 263 beds. The Clinical Examination Section consists of 12 staff (including 5 part-time workers): 7 in change of sample examination and 5 in charge of physiological examination. The annual number of samples is approximately 150,000. Samples collected on health checkups account for 90%. To improve the quality and service, assessment by third persons has been positively utilized in our hospital. Accreditation regarding ISO9001, ISO14001, ISO27001, privacy mark, hospital function assessment, the functional assessment of "ningen-dock"/health checkup hospitals, labor/hygiene service function assessment, and ISO15189 has been acquired. Patients may not recognize ISO. So, it must be utilized, considering that the acquisition of accreditation is not a goal but a starting point. Furthermore, cost-performance should be improved to achieve utilization-related merits. It is important to not only acquire accreditation but also help clinical staff and patients become aware of some changes/merits. Patients may consult a hospital for the following reasons: confidence in the hospital, and the staffs kind/polite attitudes. Long-term management strategies should be established without pursuing only short-term profits. I will introduce several merits of acquiring accreditation regarding ISO15189. Initially, incidental conditions for bids and appeal points include accreditation regarding ISO15189. Our corporation has participated in some competitive bids regarding health checkup business. In some companies, the bid conditions included ISO acquisition. In our hospital, clinical trials have been positively carried out. For participation in trials, hospitals must pass an institutional examination. However, ISO acquisition facilitates the preparation of

  7. Film dosimetry in conformal radiotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Danciu, C.; Proimos, B.S. [Patras Univ. (Greece). Dept. of Medical Physics

    1995-12-01

    Dosimetry, through a film sandwiched in a transverse cross-section of a solid phantom, is a method of choice in Conformal Radiotherapy because: (a) the blackness (density) of the film at each point offers a measure of the total dose received at that point, and (b) the film is easily calibrated by exposing a film strip in the same cross-section, through a stationary field. The film must therefore have the following properties: (a) it must be slow, in order not to be overexposed, even at a therapeutic dose of 200 cGy, and (b) the response of the film (density versus dose curve) must be independent of the photon energy spectrum. A few slow films were compared. It was found that the Kodak X-Omat V for therapy verification was the best choice. To investigate whether the film response was independent of the photon energy, response curves for six depths, starting from the depth of maximum dose to the depth of 25 cm, in solid phantom were derived. The vertical beam was perpendicular to the anterior surface of the phantom, which was at the distance of 100 cm from the source and the field was 15x15 cm at that distance. This procedure was repeated for photon beams emitted by a Cobalt-60 unit, two 6 MV and 15 MV Linear Accelerators, as well as a 45 MV Betatron. For each of those four different beams the film response was the same for all six depths. The results, as shown in the diagrams, are very satisfactory. The response curve under a geometry similar to that actually applied, when the film is irradiated in a transverse cross-section of the phantom, was derived. The horizontal beam was almost parallel (angle of 85) to the plane of the film. The same was repeated with the central ray parallel to the film (angle 90) and at a distance of 1.5 cm from the horizontal film. The field size was again 15x15 at the lateral entrance surface of the beam. The response curves remained the same, as when the beam was perpendicular to the films.

  8. Accredited Medical Web: an experience in Spain

    OpenAIRE

    Sarrias, R; Mayer, MA; M. Latorre

    2000-01-01

    Introduction The Official Medical College of Barcelona (OMCB) is a centennial corporation created to defend the interests of the medical profession and ensure that it adheres to ethical and scientific norms in order to offer the best healthcare services to society. The Internet has the capacity to transmit a volume of information that is both difficult to control and widely available. The OMCB emphasizes the necessity to contribute to the accreditation of medical/healthcare information diffus...

  9. Dosimetry requirements derived from the sterilization standards

    DEFF Research Database (Denmark)

    Miller, A.

    1998-01-01

    The main standards for radiation sterilization, ISO 11137 and EN 552, rest the documentation for the properly executed sterilization process on dosimetry. Both standards describe general requirements to the dosimetry system: The dose measurements must be traceable to national standards...

  10. Incompatibilities analysis in the accredited laboratory

    Directory of Open Access Journals (Sweden)

    D. Szewieczek

    2008-06-01

    Full Text Available Purpose of the presented paper aimed at motivating the necessity of the accreditation of research and standardising laboratories as factors deciding about the competitive advantage of those organisations on the European Union market.Design/methodology/approach used for the research has covered the analyses of results of internal and external audits conducted in one of Polish accredited laboratories and estimation of the incompatibilities occurred.Findings of the carried out research are as follows: number and character of incompatibilities, which are exposed during internal and external audits, reflect size of organisation, where the management system is implemented, phase of implementation as well as the time of functioning.Practical implications refers to any organisation which has quality management system implemented as well as to any accredited laboratory using internal audits as an element of continuous improvement and treating incompatibilities not as something disqualifying the investigated area, but as an supporting element. Originality/value of the presented paper belongs to the methodology comprising the usage of internal audits’ results - proved incompatibilities - as a tool for obtaining and assuring the confidence in the management system.

  11. International Standardization of the Clinical Dosimetry of Beta Radiation Brachytherapy Sources: Progress of an ISO Standard

    Science.gov (United States)

    Soares, Christopher

    2006-03-01

    In 2004 a new work item proposal (NWIP) was accepted by the International Organization for Standardization (ISO) Technical Committee 85 (TC85 -- Nuclear Energy), Subcommittee 2 (Radiation Protection) for the development of a standard for the clinical dosimetry of beta radiation sources used for brachytherapy. To develop this standard, a new Working Group (WG 22 - Ionizing Radiation Dosimetry and Protocols in Medical Applications) was formed. The standard is based on the work of an ad-hoc working group initiated by the Dosimetry task group of the Deutsches Insitiut für Normung (DIN). Initially the work was geared mainly towards the needs of intravascular brachytherapy, but with the decline of this application, more focus has been placed on the challenges of accurate dosimetry for the concave eye plaques used to treat ocular melanoma. Guidance is given for dosimetry formalisms, reference data to be used, calibrations, measurement methods, modeling, uncertainty determinations, treatment planning and reporting, and clinical quality control. The document is currently undergoing review by the ISO member bodies for acceptance as a Committee Draft (CD) with publication of the final standard expected by 2007. There are opportunities for other ISO standards for medical dosimetry within the framework of WG22.

  12. Thermoluminescence Dosimetry Applied to Radiation Protection

    DEFF Research Database (Denmark)

    Christensen, Poul; Bøtter-Jensen, Lars; Majborn, Benny

    1982-01-01

    This is a general review of the present state of the development and application of thermoluminescence dosimetry (TLD) for radiation protection purposes. A description is given of commonly used thermoluminescent dosimeters and their main dosimetric properties, e.g. energy response, dose range......, fading, and LET dependence. The applications of thermoluminescence dosimetry in routine personnel monitoring, accident dosimetry, u.v. radiation dosimetry, and environmental monitoring are discussed with particular emphasis on current problems in routine personnel monitoring. Finally, the present state...

  13. Calibration of well-type ionization chambers; Calibracao de camaras de ionizacao do tipo poco

    Energy Technology Data Exchange (ETDEWEB)

    Alves, C.F.E.; Leite, S.P.; Pires, E.J.; Magalhaes, L.A.G.; David, M.G.; Almeida, C.E. de, E-mail: cfealves@gmail.com [Universidade do Estado do Rio de Janeiro (UERJ), Rio de Janeiro, RJ (Brazil). Lab. de Ciencias Radiologicas; Di Prinzio, R. [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil)

    2015-07-01

    This paper presents the methodology developed by the Laboratorio de Ciencias Radiologicas and presently in use for determining of the calibration coefficient for well-type chambers used in the dosimetry of {sup 192}Ir high dose rate sources. Uncertainty analysis involving the calibration procedure are discussed. (author)

  14. Biological dosimetry; Dosimetria biologica

    Energy Technology Data Exchange (ETDEWEB)

    Guerrero C, C.; Arceo M, C., E-mail: citlali.guerrero@inin.gob.m [ININ, Departamento de Biologia, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)

    2010-07-01

    In the Instituto Nacional de Investigaciones Nucleares (ININ) the works to establish a laboratory of biological dosimetry were initiated in 1998, with the purpose that could assist any situation with respect to the exposition to radiation, so much of the occupational exposed personnel as of individuals not related with the handling of radio-active material. The first activity that was realized was to develop the corresponding curves in vitro of dose response for different qualities and radiation types. In the year 2000 the curve corresponding to the gamma radiation of {sup 60}Co was published and up to 2002 the curve corresponding to the X rays of 58 KeV, 120 and 250 kVp. In all the cases, the curves contain the requirements to be used in the determination of the exposition dose. At the present time the curves dose-response are developing for neutrons take place in the reactor Triga Mark III of ININ. Additionally to these activities, cases of suspicion of accidental exposition to radiation have been assisted, using in a beginning the curves published by the IAEA and, from the year 2000, the curves developed in the ININ. (Author)

  15. Fast neutron dosimetry

    International Nuclear Information System (INIS)

    During 1988--1990 the magnetic resonance dosimetry project was completed, as were the 250 MeV proton shielding measurements. The first cellular experiment using human cells in vitro at the 1 GeV electron storage ring was also accomplished. More detail may be found in DOE Report number-sign DOE/EV/60417-002 and the open literature cited in the individual progress subsections. We report Kinetic Energy Released in Matter (KERMA), factor measurements in several elements of critical importance to neutron radiation therapy and radiation protection for space habitation and exploration for neutron energies below 30 MeV. The results of this effort provide the only direct measurements of the oxygen and magnesium kerma factors above 20 MeV neutron energy, and the only measurements of the iron kerma factor above 15 MeV. They provide data of immediate relevance to neutron radiotherapy and impose strict criteria for normalizing and testing nuclear models used to calculate kerma factors at higher neutron energies

  16. Nuclear accident dosimetry intercomparison studies.

    Science.gov (United States)

    Sims, C S

    1989-09-01

    Twenty-two nuclear accident dosimetry intercomparison studies utilizing the fast-pulse Health Physics Research Reactor at the Oak Ridge National Laboratory have been conducted since 1965. These studies have provided a total of 62 different organizations a forum for discussion of criticality accident dosimetry, an opportunity to test their neutron and gamma-ray dosimetry systems under a variety of simulated criticality accident conditions, and the experience of comparing results with reference dose values as well as with the measured results obtained by others making measurements under identical conditions. Sixty-nine nuclear accidents (27 with unmoderated neutron energy spectra and 42 with eight different shielded spectra) have been simulated in the studies. Neutron doses were in the 0.2-8.5 Gy range and gamma doses in the 0.1-2.0 Gy range. A total of 2,289 dose measurements (1,311 neutron, 978 gamma) were made during the intercomparisons. The primary methods of neutron dosimetry were activation foils, thermoluminescent dosimeters, and blood sodium activation. The main methods of gamma dose measurement were thermoluminescent dosimeters, radiophotoluminescent glass, and film. About 68% of the neutron measurements met the accuracy guidelines (+/- 25%) and about 52% of the gamma measurements met the accuracy criterion (+/- 20%) for accident dosimetry. PMID:2777549

  17. Polymer gel dosimetry system for radiation therapy

    International Nuclear Information System (INIS)

    calibration function was created by fitting a polynomial to a set of dose and R2 data points. This calibration function is then applied to all other R2 maps, so that dose maps can be displayed, analysed and annotated. Isodose curves, beam profiles, and central axis depth dose curves may be created, and numerical values exported for comparison with corresponding data from treatment planning computer software. Results: Gels were irradiated by X rays in the 6-20 MV range and up to 20 MeV electron beams, using a range of field sizes and dose rates in the range 40-400 cGy/min, as well as by an unshielded high activity source of 192Ir. The results obtained from the polymer gels were compared to those calculated by a commercially available treatment planning system, and the agreement was found to be within the range of experimental error. Having established the ability of the polymer gel system to accurately determine well-established dose distributions, they were then applied to the more complex problems posed by the dynamic wedge, stereotactic radiosurgery, conformal therapy, and shielded sources. Wherever comparisons with standard methods were possible, the agreement was excellent. Conclusion: There is excellent agreement between the dose distributions predicted using standard treatment planning systems and those determined using the polymer gel method, and the clinical practical utility of MRI-based polymer gel dosimetry is thereby demonstrated

  18. Initial radiation dosimetry at Hiroshima and Nagasaki

    International Nuclear Information System (INIS)

    The dosimetry of A-bomb survivors at Hiroshima and Nagasaki is discussed in light of the new dosimetry developed in 1980 by the author. The important changes resulting from the new dosimetry are the ratios of neutron to gamma doses, particularly at Hiroshima. The implications of these changes in terms of epidemiology and radiation protection standards are discussed

  19. Quality indicators to compare accredited independent pharmacies and accredited chain pharmacies in Thailand.

    Science.gov (United States)

    Arkaravichien, Wiwat; Wongpratat, Apichaya; Lertsinudom, Sunee

    2016-08-01

    Background Quality indicators determine the quality of actual practice in reference to standard criteria. The Community Pharmacy Association (Thailand), with technical support from the International Pharmaceutical Federation, developed a tool for quality assessment and quality improvement at community pharmacies. This tool has passed validity and reliability tests, but has not yet had feasibility testing. Objective (1) To test whether this quality tool could be used in routine settings. (2) To compare quality scores between accredited independent and accredited chain pharmacies. Setting Accredited independent pharmacies and accredited chain pharmacies in the north eastern region of Thailand. Methods A cross sectional study was conducted in 34 accredited independent pharmacies and accredited chain pharmacies. Quality scores were assessed by observation and by interviewing the responsible pharmacists. Data were collected and analyzed by independent t-test and Mann-Whitney U test as appropriate. Results were plotted by histogram and spider chart. Main outcome measure Domain's assessable scores, possible maximum scores, mean and median of measured scores. Results Domain's assessable scores were close to domain's possible maximum scores. This meant that most indicators could be assessed in most pharmacies. The spider chart revealed that measured scores in the personnel, drug inventory and stocking, and patient satisfaction and health promotion domains of chain pharmacies were significantly higher than those of independent pharmacies (p pharmacies and chain pharmacies in the premise and facility or dispensing and patient care domains. Conclusion Quality indicators developed by the Community Pharmacy Association (Thailand) could be used to assess quality of practice in pharmacies in routine settings. It is revealed that the quality scores of chain pharmacies were higher than those of independent pharmacies. PMID:27118461

  20. Calibration of two 90Sr+90Y dermatological applicators

    International Nuclear Information System (INIS)

    The 90Sr+90Y applicators need to be periodically calibrated, but in Brazil the service it not offered yet. The recommended method for the calibration of this kind of applicators is the use of extrapolation chambers. An alternative method for the calibration of clinical applicators is the use of thermoluminescent dosimeters. A dosimetric method of these applicators was already developed at Instituto de Pesquisas Energeticas e Nucleares (IPEN) and several types of thermoluminescent dosimeters were studied in previous works. The aim of this work was the application of this method to calibrate two dermatological applicators. Thin CaSO4:Dy pellets, with and without 10% of graphite were utilized. The reproducibility of these pellets was studied, and calibration curves were obtained using a standard applicator calibrated at the National Institute of Standards and Technology (NIST), USA. Both applicators showed similar results. The TL materials tested showed usefulness for dosimetry and calibration of this kind of applicators. (author)

  1. Comment on 'Proton beam monitor chamber calibration'.

    Science.gov (United States)

    Palmans, Hugo; Vatnitsky, Stanislav M

    2016-09-01

    We comment on a recent article (Gomà et al 2014 Phys. Med. Biol. 59 4961-71) which compares different routes of reference dosimetry for the energy dependent beam monitor calibration in scanned proton beams. In this article, a 3% discrepancy is reported between a Faraday cup and a plane-parallel ionization chamber in the experimental determination of the number of protons per monitor unit. It is further claimed that similar discrepancies between calorimetry and ionization chamber based dosimetry indicate that [Formula: see text]-values tabulated for proton beams in IAEA TRS-398 might be overestimated. In this commentary we show, however, that this supporting argument misrepresents the evidence in the literature and that the results presented, together with published data, rather confirm that there exist unresolved problems with Faraday cup dosimetry. We also show that the comparison in terms of the number of protons gives a biased view on the uncertainty estimates for both detectors while the quantity of interest is absorbed dose to water or dose-area-product to water, even if a beam monitor is calibrated in terms of the number of protons. Gomà et al (2014 Phys. Med. Biol. 59 4961-71) also report on the discrepancy between cylindrical and plane-parallel ionization chambers and confirm experimentally that in the presence of a depth dose gradient, theoretical values of the effective point of measurement, or alternatively a gradient correction factor, account for the discrepancy. We believe this does not point to an error or shortcoming of IAEA TRS-398, which prescribes taking the centre of cylindrical ionization chambers as reference point, since it recommends reference dosimetry to be performed in the absence of a depth dose gradient. But these observations reveal that important aspects of beam monitor calibration in scanned proton beams are not addressed in IAEA TRS-398 given that those types of beams were not widely implemented at the time of its publication

  2. Report of the Intercomparison program by thermoluminescent dosimetry for Secondary Standard Dosimetry Laboratories

    International Nuclear Information System (INIS)

    In this report the results of an intercomparison program within a research coordinated program are presented. This is a third phase of the study that consisted in to evaluate the implementation of the new ICRU quantities for individual monitoring by the SSDLs, their capabilities to perform irradiations in different angles and the interpretation of the standard ISO 4370-3. This phase as well the first one was coordinated by Argentina through the Autoridad Regulatoria Nuclear that verified the performance of the participant laboratories. The SSDL of Argentina calibrated the dosimetric system to be used, and sent a set of tld dosimeters for irradiation at the SSDL or dosimetry laboratories of nine countries of latin america

  3. High sensitive radiation detector for radiology dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Valente, M.; Malano, F. [Instituto de Fisica Enrique Gaviola, Oficina 102 FaMAF - UNC, Av. Luis Medina Allende, Ciudad Universitaria, 5000 Cordoba (Argentina); Molina, W.; Vedelago, J., E-mail: valente@famac.unc.edu.ar [Laboratorio de Investigaciones e Instrumentacion en Fisica Aplicada a la Medicina e Imagenes por Rayos X, Laboratorio 448 FaMAF - UNC, Ciudad Universitaria, 5000 Cordoba (Argentina)

    2014-08-15

    Fricke solution has a wide range of applications as radiation detector and dosimetry. It is particularly appreciated in terms of relevant comparative advantages, like tissue equivalence when prepared in aqueous media like gel matrix, continuous mapping capability, dose rate recorded and incident direction independence as well as linear dose response. This work presents the development and characterization of a novel Fricke gel system, based on modified chemical compositions making possible its application in clinical radiology. Properties of standard Fricke gel dosimeter for high dose levels are used as starting point and suitable chemical modifications are introduced and carefully investigated in order to attain high resolution for low dose ranges, like those corresponding to radiology interventions. The developed Fricke gel radiation dosimeter system achieves the expected typical dose dependency, actually showing linear response in the dose range from 20 up to 4000 mGy. Systematic investigations including several chemical compositions are carried out in order to obtain a good enough dosimeter response for low dose levels. A suitable composition among those studied is selected as a good candidate for low dose level radiation dosimetry consisting on a modified Fricke solution fixed to a gel matrix containing benzoic acid along with sulfuric acid, ferrous sulfate, xylenol orange and ultra-pure reactive grade water. Dosimeter samples are prepared in standard vials for its in phantom irradiation and further characterization by spectrophotometry measuring visible light transmission and absorbance before and after irradiation. Samples are irradiated by typical kV X-ray tubes and calibrated Farmer type ionization chamber is used as reference to measure dose rates inside phantoms in at vials locations. Once sensitive material composition is already optimized, dose-response curves show significant improvement regarding overall sensitivity for low dose levels. According to

  4. Evolving Alignment in International Continuing Professional Development Accreditation.

    Science.gov (United States)

    McMahon, Graham T; Aboulsoud, Samar; Gordon, Jennifer; McKenna, Mindi; Meuser, James; Staz, Mark; Campbell, Craig M

    2016-01-01

    Several of the world's accreditation systems for continuing professional development (CPD) are evolving to encourage continuous improvement in the competence and performance of health care providers and in the organizations in which they provide patient care. Clinicians learn best when they can to choose from a diverse array of activities and formats that are relevant and meet their needs. Since choice and diversity are key to meeting clinicians' needs, several CPD accreditors have been engaging in deliberate, concerted efforts to identify a core set of principles that can serve as the basis for determining substantive equivalency between CPD accreditation systems. Substantive equivalency is intended to support the mobility of learners, allowing them to access accredited learning activities that are recognized by various CPD accreditation systems in a manner that maximizes the value of those accreditation systems, while minimizing the burden of adhering to their requirements. In this article, we propose a set of core principles that all CPD accreditation systems must express as the basis for determining substantive equivalency between CPD accreditation systems. The article will illustrate how five CPD accreditation systems (two in the USA, two in Canada, and one in Qatar), differing in focus (activity-based versus provider-based), context, and culture, express these values and metrics, and concludes by identifying the value of substantive equivalency for learners, medical regulators, and CPD accreditation systems.

  5. International accreditation of ambulatory surgical centers and medical tourism.

    Science.gov (United States)

    McGuire, Michael F

    2013-07-01

    The two forces that have driven the increase in accreditation of outpatient ambulatory surgery centers (ASC's) in the United States are reimbursement of facility fees by Medicare and commercial insurance companies, which requires either accreditation, Medicare certification, or state licensure, and state laws which mandate one of these three options. Accreditation of ASC's internationally has been driven by national requirements and by the competitive forces of "medical tourism." The three American accrediting organizations have all developed international programs to meet this increasing demand outside of the United States. PMID:23830758

  6. Evolving Alignment in International Continuing Professional Development Accreditation.

    Science.gov (United States)

    McMahon, Graham T; Aboulsoud, Samar; Gordon, Jennifer; McKenna, Mindi; Meuser, James; Staz, Mark; Campbell, Craig M

    2016-01-01

    Several of the world's accreditation systems for continuing professional development (CPD) are evolving to encourage continuous improvement in the competence and performance of health care providers and in the organizations in which they provide patient care. Clinicians learn best when they can to choose from a diverse array of activities and formats that are relevant and meet their needs. Since choice and diversity are key to meeting clinicians' needs, several CPD accreditors have been engaging in deliberate, concerted efforts to identify a core set of principles that can serve as the basis for determining substantive equivalency between CPD accreditation systems. Substantive equivalency is intended to support the mobility of learners, allowing them to access accredited learning activities that are recognized by various CPD accreditation systems in a manner that maximizes the value of those accreditation systems, while minimizing the burden of adhering to their requirements. In this article, we propose a set of core principles that all CPD accreditation systems must express as the basis for determining substantive equivalency between CPD accreditation systems. The article will illustrate how five CPD accreditation systems (two in the USA, two in Canada, and one in Qatar), differing in focus (activity-based versus provider-based), context, and culture, express these values and metrics, and concludes by identifying the value of substantive equivalency for learners, medical regulators, and CPD accreditation systems. PMID:27584065

  7. Metrology and quality of radiation therapy dosimetry of electron, photon and epithermal neutron beams

    Energy Technology Data Exchange (ETDEWEB)

    Kosunen, A

    1999-08-01

    In radiation therapy using electron and photon beams the dosimetry chain consists of several sequential phases starting by the realisation of the dose quantity in the Primary Standard Dosimetry Laboratory and ending to the calculation of the dose to a patient. A similar procedure can be described for the dosimetry of epithermal neutron beams in boron neutron capture therapy (BNCT). To achieve the required accuracy of the dose delivered to a patient the quality of all steps in the dosimetry procedure has to be considered. This work is focused on two items in the dosimetry chains: the determination of the dose in the reference conditions and the evaluation of the accuracy of dose calculation methods. The issues investigated and discussed in detail are: a)the calibration methods of plane parallel ionisation chambers used in electron beam dosimetry, (b) the specification of the critical dosimetric parameter i.e. the ratio of stopping powers for water to air, (S I ?){sup water} {sub air}, in photon beams, (c) the feasibility of the twin ionization chamber technique for dosimetry in epithermal neutron beams applied to BNCT and (d) the determination accuracy of the calculated dose distributions in phantoms in electron, photon, and epithermal neutron beams. The results demonstrate that up to a 3% improvement in the consistency of dose determinations in electron beams is achieved by the calibration of plane parallel ionisation chambers in high energy electron beams instead of calibrations in {sup 60}Co gamma beams. In photon beam dosimetry (S I ?){sup water} {sub air} can be determined with an accuracy of 0.2% using the percentage dose at the 10 cm depth, %dd(10), as a beam specifier. The use of %odd(10) requires the elimination of the electron contamination in the photon beam. By a twin ionisation chamber technique the gamma dose can be determined with uncertainty of 6% (1 standard deviation) and the total neutron dose with an uncertainty of 15 to 20% (1 standard deviation

  8. Metrology and quality of radiation therapy dosimetry of electron, photon and epithermal neutron beams

    International Nuclear Information System (INIS)

    In radiation therapy using electron and photon beams the dosimetry chain consists of several sequential phases starting by the realisation of the dose quantity in the Primary Standard Dosimetry Laboratory and ending to the calculation of the dose to a patient. A similar procedure can be described for the dosimetry of epithermal neutron beams in boron neutron capture therapy (BNCT). To achieve the required accuracy of the dose delivered to a patient the quality of all steps in the dosimetry procedure has to be considered. This work is focused on two items in the dosimetry chains: the determination of the dose in the reference conditions and the evaluation of the accuracy of dose calculation methods. The issues investigated and discussed in detail are: a)the calibration methods of plane parallel ionisation chambers used in electron beam dosimetry, (b) the specification of the critical dosimetric parameter i.e. the ratio of stopping powers for water to air, (S I ?)waterair, in photon beams, (c) the feasibility of the twin ionization chamber technique for dosimetry in epithermal neutron beams applied to BNCT and (d) the determination accuracy of the calculated dose distributions in phantoms in electron, photon, and epithermal neutron beams. The results demonstrate that up to a 3% improvement in the consistency of dose determinations in electron beams is achieved by the calibration of plane parallel ionisation chambers in high energy electron beams instead of calibrations in 60Co gamma beams. In photon beam dosimetry (S I ?)waterair can be determined with an accuracy of 0.2% using the percentage dose at the 10 cm depth, %dd(10), as a beam specifier. The use of %odd(10) requires the elimination of the electron contamination in the photon beam. By a twin ionisation chamber technique the gamma dose can be determined with uncertainty of 6% (1 standard deviation) and the total neutron dose with an uncertainty of 15 to 20% (1 standard deviation). To improve the accuracy of

  9. Dosimetry measurements at the fast neutron therapy facility in Seattle

    International Nuclear Information System (INIS)

    The fast neutron therapy facility at the University of Washington has been in routine clinical use for 25 years. 50.5 MeV protons produce neutrons in a beryllium target mounted on an isocentric gantry. Beam shaping is accomplished with a 40-leaf collimator. Dosimetry measurements for treatment planning and calibration are performed with tissue equivalent ion chambers. A layered phantom of alternating Solid Water and Plastic Water slabs has been developed for rapid dose verification measurements. The neutron field in the room has been used for radiation testing of electronic components.

  10. Information from the Dosimetry Service

    CERN Multimedia

    2006-01-01

    Please note the following opening hours of the Service: In June: Every morning from 8:30 to 12:00 In July: Mondays, Wednesdays and Fridays from 8:30 to 11:30 Closed all day on Tuesdays and Thursdays From 31st July onwards: Every morning from 8:30 to 12:00 The Service is closed in the afternoons. We should like to remind you that dosimeters cannot be sent to customers by internal mail. Short-term dosimeters (VCTs) must always be returned to the Service after use and must not be left on the racks in the experimental areas or in the secretariats. Dosimetry Service Tel 72155 Bldg. 24 E 011 Dosimetry.service@cern.ch http://cern.ch/rp-dosimetry

  11. External audit in radiotherapy dosimetry

    International Nuclear Information System (INIS)

    Quality audit forms an essential part of any comprehensive quality assurance programme. This is true in radiotherapy generally and in specific areas such as radiotherapy dosimetry. Quality audit can independently test the effectiveness of the quality system and in so doing can identify problem areas and minimize their possible consequences. Some general points concerning quality audit applied to radiotherapy are followed by specific discussion of its practical role in radiotherapy dosimetry, following its evolution from dosimetric intercomparison exercises to routine measurement-based on-going audit in the various developing audit networks both in the UK and internationally. Specific examples of methods and results are given from some of these, including the Scottish+ audit group. Quality audit in radiotherapy dosimetry is now well proven and participation by individual centres is strongly recommended. Similar audit approaches are to be encouraged in other areas of the radiotherapy process. (author)

  12. For information: Individual dosimetry service

    CERN Document Server

    2004-01-01

    The service has noticed that there are dosimeter holders who have changed their activities and thus have no longer need of dosimeter as a permanent basis in their work (persons who go rarely to the controlled areas). The reduction of persons in the regular distribution list of dosimeters will lighten the work of the service (distribution, evaluation and consolidation of doses) as well as the work of the distributors, needless to say the economical input this would have for CERN. For the persons who only need a dosimeter temporarily we would like to remind that there is a quick and simple procedure to have one immediately from the Individual Dosimetry Service. Please contact the service (dosimetry.service@cern.ch) if you do not need a dosimeter regularly. Thank you for your cooperation. http://cern.ch/rp-dosimetry

  13. Radiation dosimetry and radiation biophysics

    International Nuclear Information System (INIS)

    Radiation dosimetry and radiation biophysics are two closely integrated programs whose joint purpose is to explore the connections between the primary physical events produced by radiation and their biological consequences in cellular systems. The radiation dosimetry program includes the theoretical description of primary events and their connection with the observable biological effects. This program also is concerned with the design and measurement of physical parameters used in theory or to support biological experiments. The radiation biophysics program tests and uses the theoretical developments for experimental design, and provides information for further theoretical development through experiments on cellular systems

  14. Radiation dosimetry and radiation biophysics

    International Nuclear Information System (INIS)

    Radiation dosimetry and radiation biophysics are two closely integrated programs whose joint purpose is to explore the connections between the primary physical events produced by radiation and their biological consequences in cellular systems. The radiation dosimetry program includes the theoretical description of primary events and their connection with the observable biological effects. This program also is concerned with design and measurement of those physical parameters used in the theory or to support biological experiments. The radiation biophysics program tests and makes use of the theoretical developments for experimental design. Also, this program provides information for further theoretical development through experiments on cellular systems

  15. Information from the Dosimetry Service

    CERN Multimedia

    2006-01-01

    CERN Staff and Users can now consult their dose records for an individual or an organizational unit with HRT. Please see more information on our web page http://cern.ch/rp-dosimetry. The Dosimetry Service is open every morning from 8.30 - 12.00, and closed in the afternoons. We would like to remind you that dosimeters cannot be sent to customers by internal mail. Short-term dosimeters (VCT's) must always be returned to the Service after use and must not be left on the racks in the experimental areas or in the secretariats.

  16. Information from the Dosimetry Service

    CERN Multimedia

    2006-01-01

    CERN Staff and Users can now consult their dose records for an individual or an organizational unit with HRT. Please see more information on our web page: http://cern.ch/rp-dosimetry. The Dosimetry Service is open every morning from 8.30 to 12.00 and is closed in the afternoons. We would like to remind you that dosimeters cannot be sent to customers by internal mail. Short-term dosimeters (VCT's) must always be returned to the Service after use and must not be left on the racks in the experimental areas or in the secretariats.

  17. 77 FR 70446 - Medicare and Medicaid Programs; Approval of the American Association for Accreditation of...

    Science.gov (United States)

    2012-11-26

    ... American Association for Accreditation of Ambulatory Surgery Facilities (AAAASF) for Continuing CMS Approval of Its Ambulatory Surgical Center Accreditation Program AGENCY: Centers for Medicare & Medicaid... Association for Accreditation of Ambulatory Surgery Facilities (AAAASF) for continued recognition as...

  18. 76 FR 63612 - National Committee on Foreign Medical Education and Accreditation Meeting

    Science.gov (United States)

    2011-10-13

    ... National Committee on Foreign Medical Education and Accreditation Meeting AGENCY: National Committee on Foreign Medical Education and Accreditation, Office of Postsecondary Education, U.S. Department of... of the National Committee on Foreign Medical Education and Accreditation (NCFMEA). For each...

  19. Measuring up to the challenges of the 21st century. An international evaluation of the Centre for Metrology and Accreditation

    Energy Technology Data Exchange (ETDEWEB)

    Clapman, P.; Kaarls, R.; Temmes, M.

    1997-04-01

    The international evaluation of the Centre for Metrology and Accreditation (MIKES) is part of the process in which all relevant industrial and technology policy measures and organizations under the auspices of the Ministry of Trade and Industry (MTI) are being evaluated with the aim of improving their effectiveness. The overall conclusion of the evaluation is that MIKES is serving the country well. An effective national measurement system (FINMET) is being maintained which provides a wide range of calibration services covering most of the nation`s needs. The accreditation service (FINAS) is now well established, is operating effectively, and has good prospects for growth. The evaluators present, however, a number of proposals (including 33 specific recommendations) where they feel that the metrology and accreditation arrangements could be better-suited to meet future national and international challenges. According to the recommendations the Finnish quality policy framework should be developed in a consistent way. There is a need of a comprehensive governmental quality policy statement upon which the inter-ministry coordination and harmonization of various conformity assessment activities can be based. MIKES should retain its current status as an agency within MTI. The national measurement system should be more centralised and a new purpose-built national standards laboratory should be procured. The responsibility for legal metrology should be transferred to MIKES. The terms of reference and membership of Advisory Committee for Metrology, as well as the Advisory Committee for Accreditation should be revised to ensure wider representation of all relevant, and especially industrial interests

  20. Development of a quality assurance program for ionizing radiation secondary calibration laboratories

    Energy Technology Data Exchange (ETDEWEB)

    Heaton, H.T. II; Taylor, A.R. Jr. [Center for Devices and Radiological Health, Rockville, MD (United States)

    1993-12-31

    For calibration laboratories, routine calibrations of instruments meeting stated accuracy goals are important. One method of achieving the accuracy goals is to establish and follow a quality assurance program designed to monitor all aspects of the calibration program and to provide the appropriate feedback mechanism if adjustments are needed. In the United States there are a number of organizations with laboratory accreditation programs. All existing accreditation programs require that the laboratory implement a quality assurance program with essentially the same elements in all of these programs. Collectively, these elements have been designated as a Measurement Quality Assurance (MQA) program. This paper will briefly discuss the interrelationship of the elements of an MQA program. Using the Center for Devices and Radiological Health (CDRH) X-ray Calibration Laboratory (XCL) as an example, it will focus on setting up a quality control program for the equipment in a Secondary Calibration Laboratory.

  1. Higher Education Community Engagement and Accreditation: Activating Engagement through Innovative Accreditation Strategies

    Science.gov (United States)

    Sandmann, Lorilee R.; Williams, Julie E.; Abrams, Eleanor D.

    2009-01-01

    Convergence is occurring between external demands placed on U.S. higher education institutions, such as those from state and federal governments for greater accountability, and calls for higher education's recommitment to public purposes. One important example of this convergence is the redesign of accreditation processes and standards. Because of…

  2. Reactor neutron dosimetry

    International Nuclear Information System (INIS)

    An analysis of requirements and possibilities for experimental neutron spectrum determination during the reactor pressure vessel surveil lance programme is given. Fast neutron spectrum and neutron dose rate were measured in the Fast neutron irradiation facility of our TRIGA reactor. It was shown that the facility can be used for calibration of neutron dosimeters and for irradiation of samples sensitive to neutron radiation. The investigation of the unfolding algorithm ITER was continued. Based on this investigations are two specialized unfolding program packages ITERAD and ITERGS written this year. They are able to unfold data from activation detectors and NaI(T1) gamma spectrometer respectively

  3. Dosimeter calibration facilities and methods at the Radiation Measurement Laboratory of the Centre d'etudes nucleaires, Grenoble

    International Nuclear Information System (INIS)

    At the Centre d'etudes nucleaires, Grenoble, the Radiation Measurement Laboratory, which forms part of the Environmental Protection and Research Department, serves the entire Centre for purposes of dosimetry and the calibration of dose meters. The needs of radiation protection are such that one must have facilities for checking periodically the calibration of radiation-monitoring instruments and developing special dosimetry techniques. It was thought a good idea to arrange for the dosimetry and radiation protection team to assist other groups working at the Centre - in particular, the staff of the biology and radiobiology laboratories - and also bodies outside the framework of the French Commissariat a l'energie atomique. Thus, technical collaboration has been established with, for example, Grenoble's Centre hospitalier universitaire (university clinic), which makes use of the facilities and skills available at the Radiation Measurement Laboratory for solving special dosimetry problems. With the Laboratory's facilities it is possible to calibrate dose meters for gamma, beta and neutron measurements

  4. In aqua vivo EPID dosimetry

    NARCIS (Netherlands)

    Wendling, M.; McDermott, L.N.; Mans, A.; Olaciregui-Ruiz, I.; Pecharroman-Gallego, R.; Sonke, J.J.; Stroom, J.; Herk, M. van; Mijnheer, B.J.

    2012-01-01

    PURPOSE: At the Netherlands Cancer Institute--Antoni van Leeuwenhoek Hospital in vivo dosimetry using an electronic portal imaging device (EPID) has been implemented for almost all high-energy photon treatments of cancer with curative intent. Lung cancer treatments were initially excluded, because t

  5. Portal dosimetry in wedged beams

    NARCIS (Netherlands)

    Spreeuw, H.; Rozendaal, R.; Camargo, P.; Mans, A.; Wendling, M.; Olaciregui-Ruiz, I.; Sonke, J.J.; Herk, M. van; Mijnheer, B.

    2015-01-01

    Portal dosimetry using electronic portal imaging devices (EPIDs) is often applied to verify high-energy photon beam treatments. Due to the change in photon energy spectrum, the resulting dose values are, however, not very accurate in the case of wedged beams if the pixel-to-dose conversion for the s

  6. Plutonium worker dosimetry.

    Science.gov (United States)

    Birchall, Alan; Puncher, M; Harrison, J; Riddell, A; Bailey, M R; Khokryakov, V; Romanov, S

    2010-05-01

    Epidemiological studies of the relationship between risk and internal exposure to plutonium are clearly reliant on the dose estimates used. The International Commission on Radiological Protection (ICRP) is currently reviewing the latest scientific information available on biokinetic models and dosimetry, and it is likely that a number of changes to the existing models will be recommended. The effect of certain changes, particularly to the ICRP model of the respiratory tract, has been investigated for inhaled forms of (239)Pu and uncertainties have also been assessed. Notable effects of possible changes to respiratory tract model assumptions are (1) a reduction in the absorbed dose to target cells in the airways, if changes under consideration are made to the slow clearing fraction and (2) a doubling of absorbed dose to the alveolar region for insoluble forms, if evidence of longer retention times is taken into account. An important factor influencing doses for moderately soluble forms of (239)Pu is the extent of binding of dissolved plutonium to lung tissues and assumptions regarding the extent of binding in the airways. Uncertainty analyses have been performed with prior distributions chosen for application in epidemiological studies. The resulting distributions for dose per unit intake were lognormal with geometric standard deviations of 2.3 and 2.6 for nitrates and oxides, respectively. The wide ranges were due largely to consideration of results for a range of experimental data for the solubility of different forms of nitrate and oxides. The medians of these distributions were a factor of three times higher than calculated using current default ICRP parameter values. For nitrates, this was due to the assumption of a bound fraction, and for oxides due mainly to the assumption of slower alveolar clearance. This study highlights areas where more research is needed to reduce biokinetic uncertainties, including more accurate determination of particle transport rates

  7. The personal dosimetry in Mexico

    International Nuclear Information System (INIS)

    The Personal Dosimetry in Mexico, has an approximately 30 year-old history; and it had been and it is at the moment, one of the more important resources with which the personnel that works with ionizing radiation sources counts for its protection. The Personal Dosimetry begins with the film dosimetry, technique that even continues being used at the present time by some users, and the main reason of its use is for economic reasons. At the moment this technique, it has been surpassed, by the Thermoluminescent dosimetry, which has taken a lot of peak, mainly by the technological development with which it is counted at the present time; what has given as a result that this technique becomes tip technology; that supported in the characteristic of the used materials, as the handling and processing of the information associated with the new PC, digitizer cards, software etc, what has allowed increases it potential. In this work the current necessities of the market are presented as well as an analysis of the future real necessities in Mexico, at national level, the companies that provide this service and that they spread to satisfy this necessity of the market, including the different used technologies are also mentioned. The application ranges, at the same time, of the advantages and disadvantages of the different systems of Personal Dosimetry in the market. The companies that at the moment provide the service of Personal Dosimetry, its use materials and equipment in indistinct form, for the monitoring of gamma radiation, beta particles, different qualities of x-ray radiation, and sometimes neutrons. The monitoring of the exposed personnel at the diverse sources of ionizing radiation mentioned is carried out in many occasions without having with the materials (detectors), neither the appropriate infrastructure and therefore without the quality control that guarantees a correct evaluation of the dose equivalent, as a result of the exposure to the ionizing radiations; it

  8. Higher education accreditation in the Philippines: A literature review

    Directory of Open Access Journals (Sweden)

    Ching, Gregory S.

    2013-01-01

    Full Text Available This paper discusses the historical perspective, rationales, implications, different types and processes that undergoes in the accreditation of private higher educational institutions in the Philippines. As a result, higher education accreditation in the Philippines is centered on four key result areas, namely: quality of teaching and research, support for students, relations with the community, and management of resources.

  9. Accreditation of Health Educational Programs. Part 1: Staff Working Papers.

    Science.gov (United States)

    Study of Accreditation of Selected Health Educational Programs, Washington, DC.

    This publication contains the first set of working papers concerned with structure, financing, research, and expansion as they relate to the accreditation of health education programs conducted by professional agencies. Texts of these papers are included: (1) "Historical Introduction to Accreditation of Health Educational Programs" by W.K. Selden,…

  10. Primary Medical Care Provider Accreditation (PMCPA): pilot evaluation.

    NARCIS (Netherlands)

    Campbell, S.M.; Chauhan, U.; Lester, H.

    2010-01-01

    BACKGROUND: While practice-level or team accreditation is not new to primary care in the UK and there are organisational indicators in the Quality and Outcomes Framework (QOF) organisational domain, there is no universal system of accreditation of the quality of organisational aspects of care in the

  11. Reputation Cycles: The Value of Accreditation for Undergraduate Journalism Programs

    Science.gov (United States)

    Blom, Robin; Davenport, Lucinda D.; Bowe, Brian J.

    2012-01-01

    Accreditation is among various outside influences when developing an ideal journalism curriculum. The value of journalism accreditation standards for undergraduate programs has been studied and is still debated. This study discovers views of opinion leaders in U.S. journalism programs, as surveyed program directors give reasons for being…

  12. AACSB Accreditation and Possible Unintended Consequences: A Deming View

    Science.gov (United States)

    Stepanovich, Paul; Mueller, James; Benson, Dan

    2014-01-01

    The AACSB accreditation process reflects basic quality principles, providing standards and a process for feedback for continuous improvement. However, implementation can lead to unintended negative consequences. The literature shows that while institutionalism and critical theory have been used as a theoretical base for evaluating accreditation,…

  13. Accreditation of Agricultural Engineering University studies in Portugal

    OpenAIRE

    Cruz, Vasco Fitas; Silva, Luis Leopoldo

    2005-01-01

    The paper describes the evaluation and accreditation processes of high studies study programs in Portugal; the process of Quality evaluation of new study courses; Nature of engineering profession; EurAgEng recognition; Recognition process implementation;Elements for the recognition process; Some aspects to assure curricula comparability; European Accreditation Process.

  14. Accreditation and improvement in process quality: A nationwide study

    DEFF Research Database (Denmark)

    Bie Bogh, Søren; Hollnagel, Erik; Johnsen, Søren P;

    -accredited hospitals for neither stroke (P = 0.55), heart failure (P=0.88), bleeding ulcer (P=0.67) and perforated ulcer (P =0.16). Non-accredited hospitals performed better in the study period regarding stroke (Ptable 1), whereas no clear differences were found for heart failure, bleeding ulcer or perforated...

  15. Shaping Performance: Do International Accreditations and Quality Management Really Help?

    Science.gov (United States)

    Nigsch, Stefano; Schenker-Wicki, Andrea

    2013-01-01

    In recent years, international accreditations have become an important form of quality management for business schools all over the world. However, given their high costs and the risk of increasing bureaucratisation and control, accreditations remain highly disputed in academia. This paper uses quantitative data to assess whether accreditations…

  16. Rehabilitation Counselor Education Accreditation: History, Structure, and Evolution

    Science.gov (United States)

    Shaw, Linda R.; Kuehn, Marvin D.

    2009-01-01

    This review examines some of the critical factors that influenced the evolution of rehabilitation counselor education accreditation. The article discusses the history and structure of the accreditation process and the activities that have occurred to maintain the relevancy and viability of the process. Major issues that the Council on…

  17. Policy Priorities for Accreditation Put Quality College Learning at Risk

    Science.gov (United States)

    Schneider, Carol Geary

    2016-01-01

    Ensuring the quality of college learning is, beyond doubt, the most important responsibility of higher education accreditation. Yet, almost no one currently thinks that accreditation, especially at the institutional level, is what it should be for twenty-first-century students and institutions of higher education. In this article, the author…

  18. 76 FR 81793 - Net Worth Standard for Accredited Investors

    Science.gov (United States)

    2011-12-29

    ... COMMISSION 17 CFR Parts 230, 239, 270, and 275 RIN 3235-AK90 Net Worth Standard for Accredited Investors... basis of having a net worth in excess of $1 million. This change to the net worth standard was effective.... Discussion A. Net Worth Standard for Accredited Investors (1) Overview of the Amended Rules (2) Treatment...

  19. Accreditation of Medical Education in China: Accomplishments and Challenges

    Science.gov (United States)

    Wang, Qing

    2014-01-01

    As an external review mechanism, accreditation has played a positive global role in quality assurance and promotion of educational reform. Accreditation systems for medical education have been developed in more than 100 countries including China. In the past decade, Chinese standards for basic medical education have been issued together with…

  20. COAMFTE Accreditation and California MFT Licensing Exam Success

    Science.gov (United States)

    Caldwell, Benjamin E.; Kunker, Shelly A.; Brown, Stephen W.; Saiki, Dustin Y.

    2011-01-01

    Professional accreditation of graduate programs in marital and family therapy (MFT) is intended to ensure the strength of the education students receive. However, there is great difficulty in assessing the real-world impact of accreditation on students. Only one measure is applied consistently to graduates of all MFT programs, regardless of…

  1. ESR dosimetry: achievements and challenges

    Energy Technology Data Exchange (ETDEWEB)

    Baffa, O., E-mail: baffa@usp.br [Universidade de Sao Paulo, Departamento de Fisica, Av. Bandeirantes 3900, 14040-901 Ribeirao Preto, Sao Paulo (Brazil)

    2015-10-15

    Full text: Electron Spin Resonance (ESR), also known as Electron Paramagnetic Resonance (EPR) and more recently as Electron Magnetic Resonance (Emr), is a spectroscopy technique able to detect unpaired electrons such as those created by the interaction ionizing radiation with matter. When the unpaired electrons created by ionizing radiation are stable over some reasonable time, ESR can be used to measure the radiation dose deposited in the material under study. In principle, any insulating material that satisfies this requisite can be used as a dosimeter. ESR has been used in retrospective dosimetry in case of radiological accidents using natural constituents of human body such as teeth, bones and nails as well as fortuitous materials as sugar, sweeteners and plastics. When using teeth the typical detected dose is 0.5 Gy for, for X-Band spectrometers (9 GHz) and even lower doses if higher frequency spectrometers are used. Clinical dosimetry is another area of potential use of this dosimetric modality. In this application the amino acid alanine has been proposed and being used. Alanine dosimeters are very easy to prepare and require no complicated treatments for use. Alanine/ESR dosimetry satisfies many of the required properties for clinical applications such as water equivalent composition, independence of response for the energy range used in therapy and high precision. Other organic materials such as ammonium tartrate are being investigated to increase the sensitivity of ESR for clinical applications. Finally, industrial applications can also benefit from this dosimetry. The challenges to expand applications, the number of users and research groups of ESR dosimetry will be discussed. (Author)

  2. Preliminary evaluation of a Neutron Calibration Laboratory

    Energy Technology Data Exchange (ETDEWEB)

    Alvarenga, Talysson S.; Neves, Lucio P.; Perini, Ana P.; Sanches, Matias P.; Mitake, Malvina B.; Caldas, Linda V.E., E-mail: talvarenga@ipen.br, E-mail: lpneves@ipen.br, E-mail: aperini@ipen.br, E-mail: msanches@ipen.br, E-mail: mbmitake@ipen.br, E-mail: lcaldas@ipen.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Federico, Claudio A., E-mail: claudiofederico@ieav.cta.br [Instituto de Estudos Avancados (IEAv/DCTA), Sao Jose dos Campos, SP (Brazil). Dept. de Ciencia e Tecnologia Aeroespacial

    2013-07-01

    In the past few years, Brazil and several other countries in Latin America have experimented a great demand for the calibration of neutron detectors, mainly due to the increase in oil prospection and extraction. The only laboratory for calibration of neutron detectors in Brazil is localized at the Institute for Radioprotection and Dosimetry (IRD/CNEN), Rio de Janeiro, which is part of the IAEA SSDL network. This laboratory is the national standard laboratory in Brazil. With the increase in the demand for the calibration of neutron detectors, there is a need for another calibration services. In this context, the Calibration Laboratory of IPEN/CNEN, Sao Paulo, which already offers calibration services of radiation detectors with standard X, gamma, beta and alpha beams, has recently projected a new calibration laboratory for neutron detectors. In this work, the ambient equivalent dose rate (H⁎(10)) was evaluated in several positions inside and around this laboratory, using Monte Carlo simulation (MCNP5 code), in order to verify the adequateness of the shielding. The obtained results showed that the shielding is effective, and that this is a low-cost methodology to improve the safety of the workers and evaluate the total staff workload. (author)

  3. Accreditation and participatory design in the healthcare sector

    DEFF Research Database (Denmark)

    Simonsen, Jesper; Scheuer, John Damm; Hertzum, Morten

    2015-01-01

    parts of the accreditation process and fit well with clinical evidence-based thinking. We describe and compare effects-driven IT development with accreditation, in terms of the Danish Quality Model which is used throughout the Danish healthcare sector, and we discuss the prospects and challenges......We revisit the role of participatory design approaches in the light of the accreditation regime currently imposed on the Danish healthcare sector. We describe effects-driven IT development as an instrument supporting sustained participatory design. Effects-driven IT development includes specifying......, realizing, and measuring the effects from using an information technology. This approach aligns with much of the logic in accreditation but is distinguished by its focus on effects, whereas current accreditation approaches focus on processes. Thereby, effects-driven IT development might support challenging...

  4. Calibrations of pocket dosemeters using a comparison method

    International Nuclear Information System (INIS)

    This monograph is dedicated mainly to the calibration of pocket dosemeters. Various types of radiation sources used in hospitals and different radiation detectors with emphasis on ionization chambers are briefly presented. Calibration methods based on the use of a reference dosemeter were developed to calibrate all pocket dosemeters existing at the Radiation Physics and Metrology Laboratory. Some of these dosemeters were used in personnel dosimetry at hospitals. Moreover, a study was realized about factors that affect the measurements with pocket dosemeters in the long term, such as discharges due to cosmic radiation. A DBASE IV program was developed to store the information included in the hospital's registry

  5. Quality assurance programme in the external dosimetry laboratory of the center for radiation protection of Cuba

    International Nuclear Information System (INIS)

    Full text: The External Dosimetry Laboratory (LDE) of the Center for Radiation Protection and Hygiene (CPHR) is the central dosimetry service of the country, which assures the individual monitoring on a monthly basis of about 5500 workers in 1000 establishments and maintains the National Dose Registry System. A project for upgrading the personal dosimetry service was started in 2000. In March 2001 film dosimeters began to gradually replaced by TLDs, the latter becoming the only official dosimeter a year later. In order to maintain quality and ensure compliance with regulations and international recommendations, a quality assurance culture has been introduced into the laboratory and quality requirements have been developed and implemented in the daily work through the establishment of a quality assurance (QA) programme. The design of the QA programme is based on the criteria contained in the ISO, on the recommendations of the IEC, and IAEA, as well as on regulations of the National Accreditation Council. The present study deals with the structure of the Laboratory's QA programme and includes documentation, organizational and managerial aspects as well as the procedures that are implemented. The experience gained during the years of the use of the QA programme is also described. (author)

  6. Personnel neutron dose assessment upgrade: Volume 1, Personnel neutron dosimetry assessment: [Final report

    International Nuclear Information System (INIS)

    This report provides guidance on the characteristics, use, and calibration criteria for personnel neutron dosimeters. The report is applicable for neutrons with energies ranging from thermal to less than 20 MeV. Background for general neutron dosimetry requirements is provided, as is relevant federal regulations and other standards. The characteristics of personnel neutron dosimeters are discussed, with particular attention paid to passive neutron dosimetry systems. Two of the systems discussed are used at DOE and DOE-contractor facilities (nuclear track emulsion and thermoluminescent-albedo) and another (the combination TLD/TED) was recently developed. Topics discussed in the field applications of these dosimeters include their theory of operation, their processing, readout, and interpretation, and their advantages and disadvantages for field use. The procedures required for occupational neutron dosimetry are discussed, including radiation monitoring and the wearing of dosimeters, their exchange periods, dose equivalent evaluations, and the documenting of neutron exposures. The coverage of dosimeter testing, maintenance, and calibration includes guidance on the selection of calibration sources, the effects of irradiation geometries, lower limits of detectability, fading, frequency of calibration, spectrometry, and quality control. 49 refs., 6 figs., 8 tabs

  7. International symposium on standards and codes of practice in medical radiation dosimetry. Book of extended synopses

    International Nuclear Information System (INIS)

    The development of radiation measurement standards by National Metrology Institutes (NMIs) and their dissemination to Secondary Standard Dosimetry Laboratories (SSDLs), cancer therapy centres and hospitals represent essential aspects of the radiation dosimetry measurement chain. Although the demands for accuracy in radiotherapy initiated the establishment of such measurement chains, similar traceable dosimetry procedures have been implemented, or are being developed, in other areas of radiation medicine (e.g. diagnostic radiology and nuclear medicine), in radiation protection and in industrial applications of radiation. In the past few years the development of primary standards of absorbed dose to water in 60Co for radiotherapy dosimetry has made direct calibrations in terms of absorbed dose to water available in many countries for the first time. Some laboratories have extended the development of these standards to high energy photon and electron beams and to low and medium energy x-ray beams. Other countries, however, still base their dosimetry for radiotherapy on air kerma standards. Dosimetry for conventional external beam radiotherapy was probably the field where standardized procedures adopted by medical physicists at hospitals were developed first. Those were related to exposure and air kerma standards. The recent development of Codes of Practice (or protocols) based on the concept of absorbed dose to water has led to changes in calibration procedures at hospitals. The International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water (TRS 398) was sponsored by the International Atomic Energy Agency (IAEA), World Health Organization (WHO), Pan-American Health Organization (PAHO) and the European Society for Therapeutic Radiology and Oncology (ESTRO) and is expected to be adopted in many countries worldwide. It provides recommendations for the dosimetry of all types of beams (except neutrons) used in external radiotherapy and satisfies

  8. History of dose specification in Brachytherapy: From Threshold Erythema Dose to Computational Dosimetry

    Science.gov (United States)

    Williamson, Jeffrey F.

    2006-09-01

    This paper briefly reviews the evolution of brachytherapy dosimetry from 1900 to the present. Dosimetric practices in brachytherapy fall into three distinct eras: During the era of biological dosimetry (1900-1938), radium pioneers could only specify Ra-226 and Rn-222 implants in terms of the mass of radium encapsulated within the implanted sources. Due to the high energy of its emitted gamma rays and the long range of its secondary electrons in air, free-air chambers could not be used to quantify the output of Ra-226 sources in terms of exposure. Biological dosimetry, most prominently the threshold erythema dose, gained currency as a means of intercomparing radium treatments with exposure-calibrated orthovoltage x-ray units. The classical dosimetry era (1940-1980) began with successful exposure standardization of Ra-226 sources by Bragg-Gray cavity chambers. Classical dose-computation algorithms, based upon 1-D buildup factor measurements and point-source superposition computational algorithms, were able to accommodate artificial radionuclides such as Co-60, Ir-192, and Cs-137. The quantitative dosimetry era (1980- ) arose in response to the increasing utilization of low energy K-capture radionuclides such as I-125 and Pd-103 for which classical approaches could not be expected to estimate accurate correct doses. This led to intensive development of both experimental (largely TLD-100 dosimetry) and Monte Carlo dosimetry techniques along with more accurate air-kerma strength standards. As a result of extensive benchmarking and intercomparison of these different methods, single-seed low-energy radionuclide dose distributions are now known with a total uncertainty of 3%-5%.

  9. Proceedings of the 5. symposium on neutron dosimetry. Beam dosimetry

    International Nuclear Information System (INIS)

    Proceedings of the fifth symposium on neutron dosimetry, organized at Neuherberg, 17-21 September 1984, by the Commission of the European Communities and the GSF Neuherberg, with the co-sponsorship of the US Department of Energy, Office of Health and Environmental Research. The proceedings deal with research on concepts, instruments and methods in radiological protection for neutrons and mixed neutron-gamma fields, including the generation, collection and evaluation of new dosimetric data, the derivation of relevant radiation protection quantitites, and the harmonization of experimental methods and instrumentation by intercomparison programmes. Besides radiation protection monitoring, the proceedings also report on the improvement of neutron beam dosimetry in the fields of radiobiology and radiation therapy

  10. Dosimetry for environmental radon: terradex experiment

    International Nuclear Information System (INIS)

    The dosimetry of gases like radon and thoron (222Rn and 220Rn) is important in several fields of general interest, such as radioprotection uranium mines, environmental health, house construction, geophysical research, medical therapy (i.e. radon baths), background measurements for experiments that study exotic processes, study of seismic events, since 222Rn is released in soil cavities prior to the earthquake and radiometric dating of materials. In this work we will present the Terradex project which is a system capable to perform accurate measurements of 219Rn, 222Rn and 220Rn (produced by the decay chains of 235U, 238U and 232Th respectively) gases concentration in air, water or other fluid. The instrument is based on silicon microstrip detectors inserted in a cylindrical fiducial volume, connected to the front-end data acquisition electronics and to a pneumatic system providing high vacuum. The experimental apparatus, data acquisition system, details of calibration procedures and of data analysis will also be discussed

  11. Characterization of a thermoluminiscence personnel dosimetry system

    International Nuclear Information System (INIS)

    Various tests carried out to characterize a Thermoluminiscence Personnel Dosimetry Automatic System, based on the optical heating of a multielement dosemeter are presented. The dosemeter consists of Lithium Borate (Copper) and Calcium Sulphate (Thallium) phosphors. The Dosimetric System shows some outstanding features, such as its simplicity (no aditional annealing procedures are required), its short reading cycle (160 TLD per hour and its data handling capabilities (RS-232C and Parallel Printer digital ports and four analigic outputs for Glow Curve Adquisition). The tests performed have been designed to conform with the different existing international Standards and Recommendations (ANSI: N13.11-1983; IEC:Draft 45B-1987, ISO:DP 8034-1984) The new radiological quantities (I.C.R.U.-19855) have been used for calibration. The results obtained (linearity, repeatibility, detection threshold, residue, stability of stored information, etc) show the optimum performance of this dosimetric system in its aplication to routine personnel dose monitoring. Based on the dosemeter energy discriminating response, an algorithm for dose assesment has been developed. The method allows personal dose calculations within 10% and gives valuable information on the quality and energy of incident radiation, for photons from 30 to 2000 keV and for Beta penetrating radiation (Sr/Y, U). (Author)

  12. Beta-dosimetry studies at LLNL

    International Nuclear Information System (INIS)

    This paper summarizes three beta-dosimetry studies made recently at the Lawrence Livermore National Laboratory. The first study was to determine the beta-gamma exposure rates at the Los Alamos Godiva IV Critical Assembly. The beta spectra from the assembly were evaluated using absorption curves and the beta-gamma dose-rate ratios were determined at various distances from the assembly. A comparison was made of the doses determined using two types of TLD personnel dosimeters and a film badge. The readings of an Eberline RO-7 instrument and the dose rates determined by TLDs were compared. Shielding provided by various metals, gloves, and clothing were measured. The second study was to determine the beta energy response of the Eberline RO-7 instrument based on measurements made with the PTB beta sources. This study required additional calibration points for the PTB sources which were made using extrapolation chamber measurements. The third study resulted in two techniques to determine the beta energy (E/sub max/) from the readings of this-window portable survey instruments. Both techniques are based on the readings obtained using aluminium filters. One technique is for field application, requires one filter, and provides a quick estimate of the beta energy in three energy groups: 1.5 MeV. The second technique is more complex requiring measurements with two or three filters, but gives the beta energy and the approximate shape of the beta spectrum. 9 references, 6 figures

  13. Dosimetry at the Los Alamos Critical Experiments Facility: Past, present, and future

    International Nuclear Information System (INIS)

    Although the primary reason for the existence of the Los Alamos Critical Experiments Facility is to provide basic data on the physics of systems of fissile material, the physical arrangements and ability to provide sources of radiation have led to applications for all types of radiation dosimetry. In the broad definition of radiation phenomena, the facility has provided sources to evaluate biological effects, radiation shielding and transport, and measurements of basic parameters such as the evaluation of delayed neutron parameters. Within the last 15 years, many of the radiation measurements have been directed to calibration and intercomparison of dosimetry related to nuclear criticality safety. Future plans include (1) the new applications of Godiva IV, a bare-metal pulse assembly, for dosimetry (including an evaluation of neutron and gamma-ray room return); (2) a proposal to relocate the Health Physics Research Reactor from the Oak Ridge National Laboratory to Los Alamos, which will provide the opportunity to continue the application of a primary benchmark source to radiation dosimetry; and (3) a proposal to employ SHEBA, a low-enrichment solution assembly, for accident dosimetry and evaluation

  14. Quality assurance of the treatments performed with a linear accelerator by means of in vivo dosimetry

    International Nuclear Information System (INIS)

    In vivo dosimetry by means of diode detectors has been used routinely in our hospital since 1996 to guarantee the dose administrated to patients undergoing a radiotherapy treatment. The aim of this work is to present how in vivo dosimetry was implemented in our centre and which kind of errors have been discovered and corrected. Before the implementation it has to be clear which kind of errors want to be traced, the tolerance and action level, who will perform the measurements and who will evaluate them. Once all these things are clear, the first thing to do is to choose the more appropriate type of diodes and to calibrate them. The lower the tolerance level, the more accurate the calibration has to be. At this point the training and motivation of people who will be involved is very important to succeed in implementing routine use of in vivo dosimetry. Choosing one treatment unit and one easy and frequent treatment technique is a good way of starting implementation. We started with prostate treatments. In vivo entrance and exit doses were measured and dose to the ICRU point was calculated. Nowadays in vivo dosimetry is performed in the second session of all treatments (X-rays and electrons). (author)

  15. Dosimetry by ESR spectroscopy of alanine

    International Nuclear Information System (INIS)

    Dosimetry based on electron spin resonance analysis of radiation-induced free radicals in amino acids (e.g. L-alanine) is relevant to biological dosimetry applications. Typical features are a wide dose range covering more than 5 decades (1-105Gy), energy independent response for photons above 100 keV, long-term stability of the ESR signal, and fast straightforward readout technique. Typical dosimeter samples, consisting of small pellets of microcrystalline amino acids in paraffin, are rugged, non-toxic, and insensitive to surface contaminations. Moreover, they are prepared homogeneously and inexpensively in large batches and can be evaluated repeatedly and supply archival dosimetry data. They have proven to be highly useful in various applications of radiation processing and sterilization dosimetry, food irradiation, quality control, radiation dosimetry, radiation therapy measurements, and as a reference system for dosimetry mailing intercomparisons. (author)

  16. NOTE FROM THE DOSIMETRY SERVICE

    CERN Multimedia

    2002-01-01

    During March, the Dosimetry Service will be opened from 8h30 to 12h in the morning and closed every afternoon.   We have established that many people, who are provided regularly with a personal dosimeter (film badge), have changed their activity and do not need it anymore, because they do not, or only exceptionally, enter controlled areas. If you are one of these persons, please contact the Personal Dosimeter Service (tel: 72155). There is a simplified procedure for obtaining a dosimeter if you have an immediate need for short-term visits in controlled areas. A reduction of the number of persons on the regular distribution list of dosimeters would decrease our and the distributors workload. It would also contribute to significant savings in the dosimetry, and thus CERN, budget. We thank you in advance for your understanding and for your collaboration.

  17. Challenges for academic accreditation: the UK experience

    Science.gov (United States)

    Shearman, Richard; Seddon, Deborah

    2010-08-01

    Several factors (government policy, demographic trends, employer pressure) are leading to new forms of degree programmes in UK universities. The government is strongly encouraging engagement between universities and employers. Work-based learning is increasingly found in first and second cycle programmes, along with modules designed by employers and increasing use of distance learning. Engineering faculties are playing a leading part in these developments, and the Engineering Council, the engineering professional bodies and some universities are collaborating to develop work-based learning programmes as a pathway to professional qualification. While potentially beneficial to the engineering profession, these developments pose a challenge to traditional approaches to programme accreditation. This paper explores how this system deals with these challenges and highlights the issues that will have to be addressed to ensure that the system can cope effectively with change, especially the development of individually tailored, work-based second cycle programmes, while maintaining appropriate standards and international confidence.

  18. MISTI Shielding and Dosimetry Experiment Project

    Data.gov (United States)

    National Aeronautics and Space Administration — Reliable on-orbit dosimetry is necessary for understanding effects of space radiation environments on spacecraft microelectronics performance and comparison of...

  19. 15 CFR 285.13 - Denial, suspension, revocation, or termination of accreditation.

    Science.gov (United States)

    2010-01-01

    ... revoke accreditation. (1) If a laboratory's accreditation is suspended, NVLAP shall notify the laboratory... NVLAP proposes to deny or revoke accreditation of a laboratory, NVLAP shall inform the laboratory of the... period. (2) If accreditation is revoked, the laboratory may be given the option of...

  20. Radiation dosimetry using decreasing TL intensity in a few variety of silicate minerals

    International Nuclear Information System (INIS)

    When the TLD is used for Radiation Dosimetry, the calibration curve is produced by plotting increasing TL intensity as function of dose. There are cases where the TL intensity decreases as dose increases and can be used in radiation dosimetry. Such behavior can be found in green quartz, three varieties of beryl and pink tourmaline. In all these silicate crystals we can show that if we irradiate with increasing γ-dose there is a dose Dm for which the TL intensity is maximum. Of course, Dm varies depending on the crystal and irradiated crystal with the dose Dm is stable. If we take one of these crystals and irradiate with dose Dm and we irradiate with doses from low values up to 400-500 Gy we obtain a curve of decreasing TL intensity, such curve can be used as calibration curve. (authors)

  1. Topical Review: Polymer gel dosimetry

    OpenAIRE

    Baldock, C; De Deene, Y; Doran, S.; Ibbott, G; Jirasek, A.; Lepage, M.; McAuley, K B; Oldham, M; Schreiner, L J

    2010-01-01

    Polymer gel dosimeters are fabricated from radiation sensitive chemicals which, upon irradiation, polymerize as a function of the absorbed radiation dose. These gel dosimeters, with the capacity to uniquely record the radiation dose distribution in three-dimensions (3D), have specific advantages when compared to one-dimensional dosimeters, such as ion chambers, and two-dimensional dosimeters, such as film. These advantages are particularly significant in dosimetry situations where steep dose ...

  2. Fourth international radiopharmaceutical dosimetry symposium

    International Nuclear Information System (INIS)

    The focus of the Fourth International Radiopharmaceutical Dosimetry Symposium was to explore the impact of current developments in nuclear medicine on absorbed dose calculations. This book contains the proceedings of the meeting including the edited discussion that followed the presentations. Topics that were addressed included the dosimetry associated with radiolabeled monoclonal antibodies and blood elements, ultrashort-lived radionuclides, and positron emitters. Some specific areas of discussion were variations in absorbed dose as a result of alterations in the kinetics, the influence of radioactive contaminants on dose, dose in children and in the fetus, available instrumentation and techniques for collecting the kinetic data needed for dose calculation, dosimetry requirements for the review and approval of new radiopharmaceuticals, and a comparison of the effect on the thyroid of internal versus external irradiation. New models for the urinary blader, skeleton including the active marrow, and the blood were presented. Several papers dealt with the validity of traditional ''average-organ'' dose estimates to express the dose from particulate radiation that has a short range in tissue. These problems are particularly important in the use of monoclonal antibodies and agents used to measure intracellular functions. These proceedings have been published to provide a resource volume for anyone interested in the calculation of absorbed radiation dose

  3. EPR Dosimetry - Present and Future

    Energy Technology Data Exchange (ETDEWEB)

    Regulla, D.F. [GSF - National Research Centre for Environment and Health, Institute of Radiation Protection, 85764 Neuherberg (Germany)

    1999-07-01

    In the past, IAEA has played a central role in stipulating research and development in EPR high-dose standardisation as well as in coordinating and organising international dose intercomparison programs, within the Member States of the United Nations from the mid-seventies till today. The future tasks of EPR dosimetry seem to tend towards different subjects such as bio markers, biological radiation effects, post-accident dose reconstruction in the environment, and retrospective human dosimetry. The latter may be considered a promising tool for epidemiology on the way to re-define radiation risk of man for chronicle radiation exposures, based on e.g. South Ural civil population and radiation workers. There are on-going international activities in the field of standardising high-level dosimetry by the American Standards on Testing and Materials (Astm), and by the International Organisation of Standards (ISO). The International Commission on Radiation Units and Measurements (ICRU) is considering the establishment of relevant recommendations concerning industrial radiation processing, but also human dose reconstruction. (Author)

  4. Photovoltaic module certification/laboratory accreditation criteria development

    Energy Technology Data Exchange (ETDEWEB)

    Osterwald, C.R. [National Renewable Energy Lab., Golden, CO (United States); Hammond, R.L.; Wood, B.D.; Backus, C.E.; Sears, R.L. [Arizona State Univ., Tempe, AZ (United States); Zerlaut, G.A. [SC-International Inc., Phoenix, AZ (United States); D`Aiello, R.V. [RD Associates, Tempe, AZ (United States)

    1995-04-01

    This document provides an overview of the structure and function of typical product certification/laboratory accreditation programs. The overview is followed by a model program which could serve as the basis for a photovoltaic (PV) module certification/laboratory accreditation program. The model covers quality assurance procedures for the testing laboratory and manufacturer, third-party certification and labeling, and testing requirements (performance and reliability). A 30-member Criteria Development Committee was established to guide, review, and reach a majority consensus regarding criteria for a PV certification/laboratory accreditation program. Committee members represented PV manufacturers, end users, standards and codes organizations, and testing laboratories.

  5. on interpretation teaching in light of translation accreditation tests

    Institute of Scientific and Technical Information of China (English)

    徐嘉璐

    2011-01-01

    with the development of society,the interpretation accreditation tests becomes a more popular subject.what are the features of these interpretation accreditation tests and what the teachers should learn from them in their teaching are the issues we should pay attention to.based on the requirements of three different translation accreditation tests and the status of interpretation teaching,this paper analyzes the required qualifications of interpreters and put forward suggestions for teachers in interpretation teaching.this paper aims to enhance the current level of interpretation teaching.

  6. Guiding the accreditation process utilizing an oversight committee

    International Nuclear Information System (INIS)

    The highly technical and complex training necessary for nuclear utilities plus regulatory and Institute for Nuclear Power Operations (INPO) accreditation demands are causing utilities to redefine training needs. The complexity of subject matter and tasks has caused utilities to recognize the importance of training methods. The INPO accreditation oversight committee responds to the new need to emphasize and standardize educational methods, Consolidated Edison established an INPO Accreditation Oversight Committee for its Indian Point facility. This presentation will describe the committee's purpose, composition, responsibilities, and the results achieved. The committee's formulation and responsibilities and the influence of committee members on training programs and management will be discussed

  7. Dependence of the glass badge response on the different calibration phantoms

    International Nuclear Information System (INIS)

    Chiyoda Technol Corporation provides a glass badge including the GD-450 for photon dosimetry and the WNP composed of a poly-allyldiglycol carbonate, the CR39, for neutron dosimetry. For maintenance of quality on monitoring service, it is very important to establish the dose estimation formula for a calibration phantom. In this study, we evaluated the GD-450 response for photon energy range from 10 keV to 1250 keV and the WNP response for neutron energy range from thermal to 15 MeV, both by experiment and Monte Carlo calculation. The dependence of the Glass Badge response was clarified on three different calibration phantoms. (author)

  8. Pre-evaluation of an ionization chamber for clinical radiotherapy dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Neves, Lucio P.; Perini, Ana P.; Xavier, Marcos; Caldas, Linda V.E., E-mail: mxavier@ipen.b, E-mail: lcaldas@ipen.b [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Khoury, Helen J., E-mail: khoury@ufpe.b [Universidade Federal de Pernambuco (UFPE), Recife, PE (Brazil). Dept. de Energia Nuclear

    2011-07-01

    This work presents some pre-operational tests for characterization of a new homemade ionization chamber developed at the Calibration Laboratory of Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP). This chamber was designed for use in radiotherapy dosimetry. To study the utilization of this chamber in radiotherapy, some tests were undertaken: short- and medium-term stabilities, saturation curve, recombination loss, polarity effect and leakage current. All results obtained in these tests were within the international recommendations. (author)

  9. Implementation of an intraoperative electron radiotherapy in vivo dosimetry program

    International Nuclear Information System (INIS)

    Intraoperative electron radiotherapy (IOERT) is a highly selective radiotherapy technique which aims to treat restricted anatomic volumes during oncological surgery and is now the subject of intense re-evaluation. In vivo dosimetry has been recommended for IOERT and has been identified as a risk-reduction intervention in the context of an IOERT risk analysis. Despite reports of fruitful experiences, information about in vivo dosimetry in intraoperative radiotherapy is somewhat scarce. Therefore, the aim of this paper is to report our experience in developing a program of in vivo dosimetry for IOERT, from both multidisciplinary and practical approaches, in a consistent patient series. We also report several current weaknesses. Reinforced TN-502RDM-H mobile metal oxide semiconductor field effect transistors (MOSFETs) and Gafchromic MD-55-2 films were used as a redundant in vivo treatment verification system with an Elekta Precise fixed linear accelerator for calibrations and treatments. In vivo dosimetry was performed in 45 patients in cases involving primary tumors or relapses. The most frequent primary tumors were breast (37 %) and colorectal (29 %), and local recurrences among relapses was 83 %. We made 50 attempts to measure with MOSFETs and 48 attempts to measure with films in the treatment zones. The surgical team placed both detectors with supervision from the radiation oncologist and following their instructions. The program was considered an overall success by the different professionals involved. The absorbed doses measured with MOSFETs and films were 93.8 ± 6.7 % and 97.9 ± 9.0 % (mean ± SD) respectively using a scale in which 90 % is the prescribed dose and 100 % is the maximum absorbed dose delivered by the beam. However, in 10 % of cases we experienced dosimetric problems due to detector misalignment, a situation which might be avoided with additional checks. The useful MOSFET lifetime length and the film sterilization procedure should also be

  10. Linearization of dose-response curve of the radiochromic film dosimetry system

    Energy Technology Data Exchange (ETDEWEB)

    Devic, Slobodan; Tomic, Nada; Aldelaijan, Saad; DeBlois, Francois; Seuntjens, Jan; Chan, Maria F.; Lewis, Dave [Medical Physics Unit, McGill University, Montreal, Quebec H3G 1A4 (Canada) and Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montreal, Quebec H3T 1E2 (Canada); Executive Administration for Radiation Protection and Safety Medical Devices Sector, Saudi Food and Drug Authority, Riyadh, Kingdom of Saudi Arabia 13312 (Saudi Arabia); Medical Physics Unit, McGill University, Montreal, Quebec H3G 1A4 (Canada) and Department of Radiation Oncology, SMBD Jewish General Hospital, McGill University, Montreal, Quebec H3T 1E2 (Canada); Medical Physics Unit, McGill University, Montreal, Quebec H3G 1A4 (Canada); Department of Medical Physics, Memorial Sloan-Kettering Cancer Center Basking Ridge, New Jersey 07920 (United States); Ashland Inc., Wayne, New Jersey 07470 (United States)

    2012-08-15

    Purpose: Despite numerous advantages of radiochromic film dosimeter (high spatial resolution, near tissue equivalence, low energy dependence) to measure a relative dose distribution with film, one needs to first measure an absolute dose (following previously established reference dosimetry protocol) and then convert measured absolute dose values into relative doses. In this work, we present result of our efforts to obtain a functional form that would linearize the inherently nonlinear dose-response curve of the radiochromic film dosimetry system. Methods: Functional form [{zeta}= (-1){center_dot}netOD{sup (2/3)}/ln(netOD)] was derived from calibration curves of various previously established radiochromic film dosimetry systems. In order to test the invariance of the proposed functional form with respect to the film model used we tested it with three different GAFCHROMIC Trade-Mark-Sign film models (EBT, EBT2, and EBT3) irradiated to various doses and scanned on a same scanner. For one of the film models (EBT2), we tested the invariance of the functional form to the scanner model used by scanning irradiated film pieces with three different flatbed scanner models (Epson V700, 1680, and 10000XL). To test our hypothesis that the proposed functional argument linearizes the response of the radiochromic film dosimetry system, verification tests have been performed in clinical applications: percent depth dose measurements, IMRT quality assurance (QA), and brachytherapy QA. Results: Obtained R{sup 2} values indicate that the choice of the functional form of the new argument appropriately linearizes the dose response of the radiochromic film dosimetry system we used. The linear behavior was insensitive to both film model and flatbed scanner model used. Measured PDD values using the green channel response of the GAFCHROMIC Trade-Mark-Sign EBT3 film model are well within {+-}2% window of the local relative dose value when compared to the tabulated Cobalt-60 data. It was also

  11. The role of dosimetry audits in radiotherapy quality assurance: The 8 year experience in Greek radiotherapy and brachytherapy centers

    International Nuclear Information System (INIS)

    Twenty six (26) radiotherapy (RT) centers (public and private) operate in Greece up to date (Apr. 2010), where 36 linacs and 8 Co-60 teletherapy units are being used, producing 64 photon beams (Co60, 6-23 MV) and 126 electron beams (4-21 MeV) in total. Furthermore, 7 HDR Ir192 and 2 MDR/LDR Cs137 remote afterloading brachytherapy systems operate. The Ionizing Radiation Calibration Laboratory (IRCL) of the Greek Atomic Energy Commission (GAEC) runs dosimetry audits in all Greek radiotherapy centers by means of on-site visits, in order to assess the dose accuracy, to identify and resolve problems on dosimetry, to provide intercomparisons to hospitals and disseminate the IAEA TRS 398 protocol. Additionally, the GAEC's IRCL calibrates the reference dosimetry equipment of all RT centers, in terms of absorbed dose in water at Co60 beam quality and air kerma strength at Ir192 qualities. The GAEC's dosimetry audit is a continuous process: The 1st round has been completed for the photons beams (2002 - 2006), electrons (2002 - 2008) and brachytherapy (2006-2009). The 2nd round is at the final stage for photons (2006-2010) and in progress for the rest, while a 3rd round for photons has already been initiated. The audit results for the photons 1st round have been published. This work presents the results of these audits and focuses on the improvements of RT centers' dosimetry during the successive audit rounds

  12. A feasibility study using radiochromic films for fast neutron 2D passive dosimetry

    Science.gov (United States)

    Brady, Samuel L.; Gunasingha, Rathnayaka; Yoshizumi, Terry T.; Howell, Calvin R.; Crowell, Alexander S.; Fallin, Brent; Tonchev, Anton P.; Dewhirst, Mark W.

    2010-09-01

    The objective of this paper is threefold: (1) to establish sensitivity of XRQA and EBT radiochromic films to fast neutron exposure; (2) to develop a film response to radiation dose calibration curve and (3) to investigate a two-dimensional (2D) film dosimetry technique for use in establishing an experimental setup for a radiobiological irradiation of mice and to assess the dose to the mice in this setup. The films were exposed to a 10 MeV neutron beam via the 2H(d,n)3He reaction. The XRQA film response was a factor of 1.39 greater than EBT film response to the 10 MeV neutron beam when exposed to a neutron dose of 165 cGy. A film response-to-soft tissue dose calibration function was established over a range of 0-10 Gy and had a goodness of fit of 0.9926 with the calibration data. The 2D film dosimetry technique estimated the neutron dose to the mice by measuring the dose using a mouse phantom and by placing a piece of film on the exterior of the experimental mouse setup. The film results were benchmarked using Monte Carlo and aluminum (Al) foil activation measurements. The radiochromic film, Monte Carlo and Al foil dose measurements were strongly correlated, and the film within the mouse phantom agreed to better than 7% of the externally mounted films. These results demonstrated the potential application of radiochromic films for passive 2D neutron dosimetry.

  13. 7 CFR 205.500 - Areas and duration of accreditation.

    Science.gov (United States)

    2010-01-01

    ... 205.500 Agriculture Regulations of the Department of Agriculture (Continued) AGRICULTURAL MARKETING SERVICE (Standards, Inspections, Marketing Practices), DEPARTMENT OF AGRICULTURE (CONTINUED) ORGANIC FOODS PRODUCTION ACT PROVISIONS NATIONAL ORGANIC PROGRAM Accreditation of Certifying Agents § 205.500 Areas...

  14. 7 CFR 205.501 - General requirements for accreditation.

    Science.gov (United States)

    2010-01-01

    ... MARKETING SERVICE (Standards, Inspections, Marketing Practices), DEPARTMENT OF AGRICULTURE (CONTINUED) ORGANIC FOODS PRODUCTION ACT PROVISIONS NATIONAL ORGANIC PROGRAM Accreditation of Certifying Agents § 205... certifying agent under this subpart must: (1) Have sufficient expertise in organic production or...

  15. 7 CFR 205.502 - Applying for accreditation.

    Science.gov (United States)

    2010-01-01

    ... Agriculture Regulations of the Department of Agriculture (Continued) AGRICULTURAL MARKETING SERVICE (Standards, Inspections, Marketing Practices), DEPARTMENT OF AGRICULTURE (CONTINUED) ORGANIC FOODS PRODUCTION ACT PROVISIONS NATIONAL ORGANIC PROGRAM Accreditation of Certifying Agents § 205.502 Applying for...

  16. ADN Programs Accredited by the National League for Nursing, 1974

    Science.gov (United States)

    Nursing Outlook, 1974

    1974-01-01

    The complete list of programs leading to an associate degree in nursing that are accredited by the National League for Nursing is presented, without annotation. The institutions are listed alphabetically by State. (Author/AJ)

  17. Accreditation and Participatory Design in the Health-Care Sector

    DEFF Research Database (Denmark)

    Simonsen, Jesper; Scheuer, John Damm; Hertzum, Morten

    2015-01-01

    We reconsider the role of participatory design approaches emphasizing the current context of the accreditation regime imposed on the Danish healthcare sector. We describe effects-driven IT development as an instrument supporting sustained participatory design. Effects-driven IT development includes...... specifying, realizing, and measuring effects from using an information technology. This approach aligns with much of the logic inherent in accreditation and it supports challenging parts of the accreditation process. Effects-driven IT development furthermore might support effects related to clinical evidence......-based thinking. We describe and compare effects- driven IT development with accreditation and discuss the prospects and challenges for this approach to participatory design within the healthcare domain....

  18. ACCREDITATION OF OPEN AND DISTANCE LEARNING: A Framework for Turkey

    Directory of Open Access Journals (Sweden)

    Serpil KOCDAR

    2012-07-01

    Full Text Available The purpose of this study is to develop a framework for the accreditation of higher open and distance learning (ODL programs in Turkey. The study was designed as a sequential monomethod multistrand mixed model including two strands which were both qualitative (QUAL→QUAL. In the first strand, both quantitative and qualitative data were collected through a three-round Delphi study with an expert panel consisting of 28 experts. In the second strand, qualitative data were collected via focus group interview. Based on a comprehensive literature review and the findings from the study, a framework was proposed including an initial accreditation process for new ODL programs and a re-accreditation process for ongoing programs. In addition, 35 criteria for new programs and 42 criteria for ongoing programs were developed to be used in the accreditation process.

  19. Accreditation of medical schools: the question of purpose and outcomes.

    Science.gov (United States)

    Azila, N M A; Tan, C P L

    2005-08-01

    Accreditation is a process by which official accrediting bodies evaluate institutions using a set of criteria and standards, following established procedures, to ensure a high quality of education needed to produce highly competent graduates. Additional objectives include (1) ensuring quality institutional functioning, (2) strengthening capabilities of educational institutions for service to the nation and (3) improving public confidence in medical schools. The accreditation process provides an opportunity for the institution to critically reflect upon all the aspects of its programme and the level of compliance or attainment of the requirements. The self-evaluation exercise, which identifies strengths and weaknesses, is perceived as formative. It is envisaged that eventually institutions will adopt a learning culture for curriculum development, implementation, monitoring and matching the outcomes. In conclusion, periodic accreditation activities can act as a "monitoring" system to ensure that the quality of medical education is maintained according to established standards. PMID:16315622

  20. International Federations (IF) accreditation instructions manual: Nanjing 2014

    OpenAIRE

    2014-01-01

    This accreditation instructions manual for International Federations is developed by the Nanjing Youth Olympic Games Organising Committee (NYOGOC) in accordance with the guidelines established in the Youth Olympic Games event manual by the International Olympic Committee (IOC).

  1. Virginia Tech's Cook Counseling Center receives international counseling accreditation

    OpenAIRE

    DeLauder, Rachel

    2010-01-01

    The Virginia Tech Thomas E. Cook Counseling Center has been accredited by the International Association of Counseling Services, Inc., an organization of United States, Canadian, and Australian counseling agencies based in Alexandria, Va.

  2. The accreditation process of careers in MERCOSUR: would the veterinary career in Uruguay achieve accreditation in accordance with the required standards of the Argentinian and Brazilian agencies?

    Directory of Open Access Journals (Sweden)

    José Passarini

    2015-06-01

    Full Text Available MERCOSUR created the ARCU-SUR system as a regional mechanism for accreditation of undergraduate programmes. The Veterinary career in Uruguay achieved accreditation by ARCU-SUR in 2010. As re-accreditation approaches an evaluation of the career with the Brazilian and Argentinian national standards of accreditation was simulated due to their strong influence on the definition of the regional criteria. Institutional self-assessment was updated and compared with the accreditation indicators of the Brazilian and Argentinian agencies. The idea of quality of the accreditation system in Brazil emphasizes the training of the teaching staff while in Argentina emphasizes the teaching and learning processes. The career could achieve accreditation with the Argentinian parameters but should improve aspects of the academic training of their teachers to achieve accreditation in Brazil. This work made it possible to adjust the Development Plan of the Faculty of Veterinary Medicine.

  3. SIXTH ERDA WORKSHOP ON PERSONNEL NEUTRON DOSIMETRY

    Energy Technology Data Exchange (ETDEWEB)

    Vallario, E. J.; Hankins, D. E.; Bramson, P. E.

    1977-07-11

    This workshop was the sixth of a series and was held on July 11 and 12, 1977, at the Oak Ridge National Laboratory in Oak Ridge, Tennessee. Those presenting papers at the Sixth Workshop prepared summary reports of their recent work for inclusion in this document. The reports are reproduced here as submitted by the participants, with only minor editing. This year's Workshop took a decidedly international flavor, with participants from seven countries in addition to the United States. The significance of this group's contributions has raised the possibility that the next Neutron Dosimetry Workshop may be held in Europe. Of particular interest at the Workshop was the keynote address by Dr. Harald Rossi. He commented that there is evidence that 1) accepted values of RBE for low absorbed doses of neutrons may be low by an order of magnitude or more and 2) the risk of leukemia is significant at 0.5 rad to the bone narrow. A reduction of the limit for permissible neutron exposure, which could result from consideration of this information, would necessitate major improvements in our "middle ages" neutron dosimetry. A number of participants reported conversions to thermoluminescent dosimeter (TLD) systems. This move has not been unanimous, however, as there were several reports of apparently satisfactory fission fragment, activation foil, and NTA film dosimeters. While thementionof NTA film resulted in the usual discussion of energy cut off and humidity effects, it seems the use of NTA in accelerator environments still has some merit. Discussion of fission fragment dosimeters centered around track etching techniques, which have shown some improvement. Of particular interest was Tommasino's report on the use of polycarbonate centrifuge tubes as the sensitive element. Thermally stimulated exoelectron emission (TSEE), never very popular for personnel dosimetry, has lost additional ground with the report that the neutron/gamma response ratio is much less than

  4. Developing a Model for Evaluation and Accreditation of Hospitals

    Directory of Open Access Journals (Sweden)

    AM Mosadegh Rad

    2005-10-01

    Full Text Available Background: Evaluation and accreditation of Health Care organizations has an important role in increasing the effectiveness and efficiency of these organizations and developing the quality of services provided. The current system of evaluation and accreditation of hospitals in Iran is based on structure only with out considering the context, process and output and this cause many problems for both health care providers and customers. Therefore, there is a pressing need for developing a model for health care systems’ evaluation and accreditation, which help these organizations to improve the quality of services. Method: The aim of this article is therefore: “To identify the strengths and weakness of the current system of evaluation and accreditation of hospitals in Iran, determining the ideal system of evaluation and accreditation to develop a model so that health care organizations aiming at achieving effectiveness and efficiency”. A descriptive study using a cross-sectional survey for data collection performed. All of hospital managers and hospital evaluators in Isfahan province were studied via a self administrated questionnaire (53 managers and 30 evaluators. This questionnaire specifies the respondents’ opinions about the current and ideal system of evaluation and accreditation of hospitals. Data was analyzed via SPSS11 software. Findings: The mean score of current system of evaluation and accreditation of hospitals was 3.12 0.83 and 4.41 0.34 (in a 5 scale.The mean score of structure, process and outcome were 3.10.73, 3.120.91and 3.091 in the current system of hospital evaluation respectively. From the view points of hospital managers and evaluators the differences between values of current and ideal scores of evaluation and accreditation system were statistically significant (P< 0.05. Therefore, there is a need for developing a model for hospital evaluation and accreditation. Based on these results an empirical model

  5. Answer to request on the ININ internal dosimetry; Respuesta a encuesta sobre dosimetria interna del ININ

    Energy Technology Data Exchange (ETDEWEB)

    Alfaro L, M.M. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

    1999-05-15

    In this report it is presented the reply to CNSNS asking for information about the methodology for the evaluation of the occupational dose due to internal contamination. The characteristics of the installation, type and dimensions of the shield room, construction materials, type of detecting, calibration geometries, type of used phantom, intervals of energy of the calibration, type of routine measurements, detection limit for Cs-137 and Co-60, code to carry out the analysis of the spectra, evaluation of the measurement data, whole body system type armchair with anthropomorphic phantom, whole body system of vertical scanning, distribution and location diagram of the internal dosimetry laboratory there are among the treated aspects. (Author.

  6. Quality Assessment of Family Medicine Teams Based on Accreditation Standards

    OpenAIRE

    Valjevac, Salih; Ridjanovic, Zoran; Masic, Izet

    2009-01-01

    CONFLICT OF INTEREST: NONE DECLARED In order to speed up and simplify the self assessment and external assessment process, provide better overview and access to Accreditation Standards for Family Medicine Teams and better assessment documents archiving, Agency for Healthcare Quality and Accreditation in Federation of Bosnia and Herzegovina (AKAZ) has developed self assessment and externals assessment software for family medicine teams. This article presents the development of standardized sof...

  7. Patient dosimetry and protection in diagnostic radiology

    International Nuclear Information System (INIS)

    The paper discussed the following subjects: x-ray in medicine as they represent by far the largest man-made sources of population exposure to ionizing radiation, methods of patient dosimetry, entrance surface dose per radiograph, dose-area product per examination, dosimetry for assessing risk, potential of dose reduction

  8. High dose dosimetry for radiation processing

    International Nuclear Information System (INIS)

    Radiation processing today offers various advantages in the field of sterilization of medical and pharmaceutical products, food preservation, treatment of chemical materials and a variety of other products widely used in modern society, all of which are of direct relevance to health and welfare. The safety and economic importance of radiation processing is clearly recognized. It is understood that reliable dosimetry is a key parameter for quality assurance of radiation processing and irradiated products. Furthermore, the standardization of dosimetry can provide a justification for the regulatory approval of irradiated products and form the basis of international clearance for free trade. After the initiation of the Agency's high dose standardization programme (1977), the first IAEA Symposium on High Dose Dosimetry was organized in 1984. As a result, concern as to the necessity of reliable dosimetry has greatly escalated not only in the scientific community but also in the radiation processing industry. The second International Symposium on High Dose Dosimetry for Radiation Processing was held in Vienna from 5 to 9 November, 1990, with a view to providing an international forum for the exchange of technical information on up to date developments in this particular field. The scientific programme held promises for an authoritative account of the status of high dose dosimetry throughout the world in 1990. Forty-one papers presented at the meeting discussed the development of new techniques, the improvement of reference and routine dosimetry systems, and the quality control and assurance of dosimetry. Refs, figs and tabs

  9. Studies on dosimetric tests applying source irradiation force of Cs-137 for using in chambers for calibration and TLD type dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Ribeiro, Laila Lorena X. [Pontificia Univ. Catolica de Goias (PUC-GO), Goiania, GO (Brazil). Dept. de Matematica, Fisica, Quimica e Engenharia de Alimentos; Barbosa, Rugles Cesar, E-mail: rbarbosa@cnen.gov.br [Centro Regional de Ciencias Nucleares do Centro Oeste (CRCN-CO/CNEN-GO), Abadia de Goias, GO (Brazil). Laboratorio de Radioprotecao; Correa, Rosangela S., E-mail: rcorrea@cnen.gov.br [Centro Regional de Ciencias Nucleares do Centro Oeste (CRCN-CO/CNEN-GO), Abadia de Goias, GO (Brazil). Laboratorio de Imagens e Dosimetria

    2011-07-01

    The West Central region of Brazil does not have a basic infrastructure for research, development, training programs, and personnel dosimetry education. All of them applied to environmental, industrial and medical uses. Service deployment for irradiance of TLD, via {sup 137}Cs irradiator J. L. SHEPHERD model 28-8A (444 activity GBq) in CRCN-CO, it is necessary to introduce procedures for calibration of the radiator and other procedures related to dosimetry and calibration. Such procedures should be repeated periodically, as necessary to introduce techniques that make the service of the CRCN-CO a template, and that meet all standards requirements for radioprotection and operation of dosimetry and calibration. The objective of this work was to evaluate the radiation field of Cs-137, and the automatic system which systematizes the calibration procedures attached to a system control target for the radiator/calibration of monitors, and portable dosimeters. (author)

  10. The history of European public health education accreditation in perspective

    Directory of Open Access Journals (Sweden)

    Julien D. Goodman

    2015-12-01

    Full Text Available Aim: The aim of this paper is to investigate the history of accreditation of academic public health education and understand why there is a 65 year gap between the first system in America and the uptake of accreditation in Europe. The paper intends to search for parallels and dissimilarities between the development in America and Europe and then consider if any parallels could be used for determining the future role of accreditation in Europe. Methods: The paper draws heavily upon a literature review and analysis and the examination and interpretation of primary and secondary sources. Firstly there is an exploration of the American development which is complemented by an evaluation of the developments in Europe. Results: The paper demonstrates that there are two key features required for the development of accreditation: interstate collaboration and a liberalisation or opening up of the education market. Conclusions: Since the Second World War, Europe has embraced interstate collaboration which has led to a liberalisation of certain economic markets. The future for sector based accreditation of public health education will be determined by the extent Europe pursues liberalisation and whether a competitive environment will bring into question the transparency and trust in state sponsored accreditation agencies.

  11. Comparison of calibrations of a well type ionization chamber between the IAEA and the SSDL of Finland

    International Nuclear Information System (INIS)

    Since 1996, the IAEA has maintained standards for Low Dose Rate (LDR) brachytherapy dosimetry. These standards consist of two 137Cs sources, calibrated at the National Institute of Standards and Technology (NIST), USA. As with all calibrations, maintaining our knowledge and confidence in of the standards at the highest possible level is essential. One way of verifying the quality of our calibrations is by means of a comparison with another SSDL. The purpose of this report is to describe such a comparison. A comparison was performed between the IAEA and the SSDL of Finland (STUK). A well type chamber, HDR 1000Plus (Standard Imaging, USA) was calibrated at the IAEA Dosimetry Laboratory and sent to STUK. After calibration by STUK the chamber was returned to the IAEA and a check of the chamber's response was made. This was done in order to verify that the chamber calibration had not been altered as a result of transportation

  12. Introduction of a new dosimetry system based on optically stimulated luminescence (OSL) in our personal monitoring service

    Energy Technology Data Exchange (ETDEWEB)

    Hubner, S., E-mail: stephan.huebner@helmholtz-muenchen.de [Helmholtz Zentrum Munchen, German Research Center for Environmental Health, D-80219, Munich (Georgia)

    2014-08-15

    The personal monitoring service named Auswertungsstelle is part of the Helmholtz Zentrum Munchen, a non-profit-making research center in Germany. As one of the four monitoring services in Germany, we have been a reliable partner in radiation protection for more than 60 years. With about 1.9 million dose assessments per year, we are the largest monitoring service in Europe. For dozens of years, our main dosimeter used in whole-body dosimetry has been a film dosimeter. Although its dosimetric properties are still up to date, film dosimetry won.t be a sustainable technique for the use in monitoring services. Therefore, a project with the objective of investigating alternative dosimetric materials and methods was launched in the late 1990 at the Helmholtz Zentrum Munchen. Based on this research work, the use of Be O as an OSL dosimeter was studied by the radiation physics group of the Tu Dresden, by order and on account of the Auswertungsstelle at the Helmholtz Zentrum Munchen. It was shown, that ceramic Be O features promising dosimetric properties, making Be O detectors particularly suitable for being used in all applications in whole-body dosimetry measuring photons. Ceramic Be O material has an excellent resistance to environmental influences. The Be O chips are almost tissue equivalent. Therefore, these detectors show low photon energy dependence. A new personal dosimetry system based on the OSL dosimetry of Be O was developed. Applying this system, the Auswertungsstelle offers OSL-dosimeters for official monitoring of the Personal Dose Equivalent Hp(10) since 2011. This OSL-System is accredited according to DIN IEC 62387 and we obtained the corresponding type approval by the Ptb, the national metrology institute in Germany. Sophisticated logistics was developed and installed. High degree of automation was achieved by robots for dosimeter assembly and machines for packing, labelling and unpacking of the dosimeters. To become a sustainable dosimetry system not only

  13. Investigation of the applicability of a special parallel-plate ionization chamber for x-ray beam dosimetry

    Science.gov (United States)

    Perini, Ana P.; P. Neves, Lucio; E. Caldas, Linda V.

    2014-02-01

    Diagnostic x-rays are the greatest source of exposition to ionizing radiation of the population worldwide. In order to obtain accurate and lower-cost dosimeters for quality control assurance of medical x-ray facilities, a special ionization chamber was designed at the Calibration Laboratory of the IPEN, for dosimetry in diagnostic radiology beams. For the chamber characterization some tests were undertaken. Monte Carlo simulations were proposed to evaluate the distribution of the deposited energy in the sensitive volume of the ionization chamber and the collecting electrode effect on the chamber response. According to the obtained results, this special ionization chamber presents potential use for dosimetry of conventional diagnostic radiology beams.

  14. Obtaining accreditation by the pharmacy compounding accreditation board, part 3: developing a program of qualtity assurance and continuous qualtiy improvement.

    Science.gov (United States)

    Cabaleiro, Joe

    2008-01-01

    Before a compounding pharmacy can receive accreditation from the Pharmacy Compounding Accreditation Board, the pharmacy must show evidence of both quality assurance activities and continuous quality improvement activities. Although quality assurance data gathering and monitoring can be integrated into pharmacy activities fairly easily, the coninuous quality improvement program may take a little more time and effort to implement . Before integrating these programs, compounding pharmacists must have a complete understanding of the differences between these two programs. Even if accreditation with the Pharmacy Compounding Accreditation Board is not being considered, it is important that compounding pharmacies have these two programs implemented. In the long run, it will pay off in higher quality services, error prevention, and perhaps greater efficiency of pharmacy operations. PMID:23969712

  15. Intercomparison of dispersed radiation readings among film dosimetry, electronic and OSL with X-rays for low dose; Intercomparacion de lecturas de radiacion dispersa entre dosimetria film, electronica y OSL con rayos X para dosis bajas

    Energy Technology Data Exchange (ETDEWEB)

    Andisco, D. [Universidad de Buenos Aires, Facultad de Medicina, Paraguay 2155, C1121AAA Buenos Aires (Argentina); Blanco, S. [CONICET, Saavedra 15, C1083ACA Buenos Aires (Argentina); Bourel, V.; Schmidt, L. [Universidad Favaloro, Facultad de Ciencias e Ingenieria, Solis 453, C1078AAI, Buenos Aires (Argentina); Di Risio, C., E-mail: dandisco@fmed.uba.ar [Universidad de Belgrano, Facultad de Ingenieria, Zabala 1837, C1426DQG, Buenos Aires (Argentina)

    2014-08-15

    One of the personal dosimetry methods more used for several decades is the dosimetry type film, characterized to possess readings with certain margin of trust. Today other methods exist that many times are presupposed more reliable due to the nature of the detection like the electronic dosimeters or the OSL (Optically Stimulated Luminescence) dosimetry. With the purpose of comparing different methods and to can determining the existent differences among each method has been carried out an intercomparison assay. The different dosimeters have been exposed to dispersed radiation generated by a Hemodynamics equipment of the type -arch in C- and a dispersing system of the primary beam. Film dosimeters have been used; OSL (In Light), OSL (Nano Dots) and Electronic with the purpose of knowing and to valorize the existent differences among its readings. Always, the intercomparison exercises have demonstrated to be an useful tool when establishing the measurement capacity and the quality of the results emitted by the laboratories of personal dosimetry services. Also, this type of assays allows obtaining quality indicators of the laboratory performance and they are habitual part of the procedures for accreditation of the same ones. The Optically Stimulated Luminescence is a technology that has grown in Argentina so much in the area of personal dosimetry as in dosimetry in vivo (radiotherapy area). In this intercomparison study, the answers corresponding to each technology were looked for oneself irradiation of the disperse type, that is to say, of very low energy. (Author)

  16. Dosimetry characterization of the commercial CaF{sub 2} for beta radiation of {sup 90}Sr + {sup 90}Y; Caracterizacao dosimetrica de CaF{sub 2} comercial para radiacao beta de {sup 90}Sr + {sup 90}Y

    Energy Technology Data Exchange (ETDEWEB)

    Oliveira, Mercia L.; Caldas, Linda V.E. [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil)], e-mail: mlolivei@ipen.br, e-mail: lcaldas@ipen.br

    2003-07-01

    This work studies the dosimetric characteristics of the CaF{sub 2} commercial dosimetry for detection of {sup 90}Sr + {sup 90}Y beta radiation for using in the calibration of flat and concave appliers. Were determined the repetitiousness and linearity of answers of the samples, and their calibration curves.

  17. RADON PROGENY AS AN EXPERIMENTAL TOOL FOR DOSIMETRY OF NANOAEROSOLS

    Energy Technology Data Exchange (ETDEWEB)

    Ruzer, Lev; Ruzer, Lev S.; Apte, Michael G.

    2008-02-25

    The study of aerosol exposure and dosimetry measurements and related quantitation of health effects are important to the understanding of the consequences of air pollution, and are discussed widely in the scientific literature. During the last 10 years the need to correlate aerosol exposure and biological effects has become especially important due to rapid development of a new, revolutionary industry ?-- nanotechnology. Nanoproduct commerce is predicted to top $1 trillion by 2015. Quantitative assessment of aerosol particle behavior in air and in lung deposition, and dosimetry in different parts of the lung, particularly for nanoaerosols, remains poor despite several decades of study. Direct measurements on humans are still needed in order to validate the hollow cast, animal studies, and lung deposition modeling. We discuss here the use of nanoscale radon decay products as an experimental tool in the study of local deposition and lung dosimetry for nanoaerosols. The issue of the safe use of radon progeny in such measurements is discussed based on a comparison of measured exposure in 3 settings: general population, miners, and in a human experiment conducted at the Paul Scherer Institute (PSI) in Switzerland. One of the properties of radon progeny is that they consist partly of 1 nm radioactive particles called unattached activity; having extremely small size and high diffusion coefficients, these particles can be potentially useful as radioactive tracers in the study of nanometer-sized aerosols. We present a theoretical and experimental study of the correlation between the unattached activity and aerosol particle surface area, together with a description of its calibration and method for measurement of the unattached fraction.

  18. Radon progeny as an experimental tool for dosimetry of nanoaerosols

    International Nuclear Information System (INIS)

    The study of aerosol exposure and dosimetry measurements and related quantitation of health effects are important to the understanding of the consequences of air pollution, and are discussed widely in the scientific literature. During the last 10 years the need to correlate aerosol exposure and biological effects has become especially important due to rapid development of a new, revolutionary industry — nano technology. Nano product commerce is predicted to top $1 trillion by 2015. Quantitative assessment of aerosol particle behavior in air and in lung deposition, and dosimetry in different parts of the lung, particularly for nano aerosols, remains poor despite several decades of study. Direct measurements on humans are still needed in order to validate the hollow cast, animal studies, and lung deposition modeling. We discuss here the use of nano scale radon decay products as an experimental tool in the study of local deposition and lung dosimetry for nano aerosols. The issue of the safe use of radon progeny in such measurements is discussed based on a comparison of measured exposure in 3 settings: general population, miners, and in a human experiment conducted at the Paul Scherer Institute (PSI) in Switzerland One of the properties of radon progeny is that they consist partly of 1 nm radioactive particles called unattached activity; having extremely small size and high diffusion coefficients, these particles can be potentially useful as radioactive tracers in the study of nanometer-sized aerosols. We present a theoretical and experimental study of the correlation between the unattached activity and aerosol particle surface area, together with a description of its calibration and method for measurement of the unattached fraction. (author)

  19. Assessment of the uncertainty and the proficiency test for accrediting KOLAS of ISO 17025 for a neutron radiography facility

    Energy Technology Data Exchange (ETDEWEB)

    Oh, H.; Sim, C.M.; Lim, I.C.; Hong, K.P.; Choi, B.H. [Korea Atomic Energy Research Inst., Daejeon (Korea)

    2004-07-01

    KOLAS(Korea of Lab Accreditation Scheme) is the charter member of ILAS (International Lab Accreditation Scheme) and APLAS (Asia Pacific Lab Accreditation Scheme), which originates from ISO 17025. KATS (Korea Agent of Technology Standard) governs the KOLAS. The KOLAS describes the basis of satisfying those issues related to a quality assurance and management system. The requirements specify an organization, the accommodation and environmental conditions, an uncertainty in the measurement and an inter-laboratory comparison or proficiency test program. The evaluation process of the requirements of certifying KOLAS for HANARO NRF has been proceeded by a neutron radiography laboratory, NRT level II course of SNT-TC-1A II is opened, with 20 persons attending for certification. An inter-laboratory comparison or proficiency test program is conducted through with Kyoto University in accordance with ASTM method for determining the imaging quality in direct thermal neutron radiographic testing (E545-91). In order to determine the uncertainty, dimensional measurements for the calibration fuel pin of the RISO using a profile project is performed with the ASTM practice for thermal neutron radiography of materials (E748-95) (orig.)

  20. Comparative study among calibration methods of clinical applicators of beta radiation

    International Nuclear Information System (INIS)

    90Sr+90Y clinical applicators are instruments used in brachytherapy procedures and they have to be periodically calibrated, according to international standards and recommendations. In this work, four calibration methods of dermatological and ophthalmic applicators were studied, comparing the results with those given by the calibration certificates of the manufacturers. The methods included the use of the standard applicator of the Calibration Laboratory (LCI), calibrated by the National Institute of Standards and Technology; an Amersham applicator (LCI) as reference; a mini-extrapolation chamber developed at LCI as an absolute standard; and thermoluminescent dosimetry. The mini-extrapolation chamber and a PTW commercial extrapolation chamber were studied in relation to their performance through quality control tests of their response, as leakage current, repeatability and reproducibility. The distribution of the depth dose in water, that presents high importance in dosimetry of clinical applicators, was determined using the mini extrapolation chamber and the thermoluminescent dosimeters. The results obtained were considered satisfactory for the both cases, and comparable to the data of the IAEA (2002) standard. Furthermore, a dosimetry postal kit was developed for the calibration of clinical applicators using the thermoluminescent technique, to be sent to clinics and hospitals, without the need of the transport of the sources to IPEN for calibration. (author)

  1. ESR dosimetry using eggshells and tooth enamel for accidental dosimetry

    International Nuclear Information System (INIS)

    The CO2- signal of eggshells showed a good dose linearity and was appropriate in the wide dose range from 1 to 10 kGy, while ESR signal of CO2- in sea and fresh water shells were saturated at a dose od below 10 kGy. The minimum detectable dose and G-value of CO2- in eggshells were estimated 0.3 Gy and 0.28, respectively. The lifetime of CO2- in eggshells could not be determined exactly because of overlapping organic signals, however it is still sufficiently long for practical use as ESR dosimeter materials. Various bird's or reptile's eggshells would be available as natural retrospective ESR dosimeter materials after nuclear accidents. Eggshells will be useful for the food irradiation dosimetry in the dose range of about a few kGy. Tooth enamel is one of the most useful dosimeter materials in public at a accident because of its high sensitivity. ESR dosimetry will replace TLD in near future if the cost of an ESR reader is further reduced . (author)

  2. Air Force neutron dosimetry program

    International Nuclear Information System (INIS)

    Approximately 1000 Air Force personnel are monitored for neutron radiation resulting from various sources at more than thirty worldwide locations. Neutron radiation spanning several orders of magnitude in energy is encountered. The Air Force currently uses albedo thermoluminescent neutron dosimeters for personnel monitoring. The energy dependence of the albedo neutron dosimeter is a current problem and the development of site specific correction factors is ongoing. A summary of data on the energy dependence is presented as well as efforts to develop algorithms for the dosimeter. An overview of current Air Force neutron dosimetry users and needs is also presented

  3. Radiation dosimetry by potassium feldspar

    Indian Academy of Sciences (India)

    Arun Pandya; S G Vaijapurkar; P K Bhatnagar

    2000-04-01

    The thermoluminescence (TL) properties of raw and annealed feldspar have been studied for their use in gamma dosimetry. The raw gamma exposed feldspar shows glow peaks at 120°C and 319°C. Gamma dose beyond 500 cGy can be measured without any significant fading even after 40 days of termination of exposure. The annealed feldspar shows a glow peak at 120°C after gamma exposure. This peak can be used to measure gamma doses beyond 25 cGy when the TL is measured after 24 h from termination of exposure.

  4. Developments in infertility counselling and its accreditation.

    Science.gov (United States)

    Monach, Jim

    2013-03-01

    Infertility counselling was placed in a unique position by the passage of the Human Fertilisation and Embryology Act 1990 and the requirement that couples being treated should be offered counselling. However professional counselling was, and largely still is, at a stage at which there was no universal agreement on the knowledge, standards or qualifications required for practice. Nevertheless, infertility counselling became the first example of counselling to be required by statute, beyond the more generalised requirement in adoption birth records access. Counselling is intended to describe skilled talking therapy offered by a professional with specific training and qualifications directed to helping individuals and couples to achieve goals they own themselves. The therapeutic intervention of counselling is primarily directed to helping clients in a stressful situation to deploy their own coping skills effectively and thus make the difficult choices inseparable from ART. Counselling outcome research consistently demonstrates the effectiveness of the sort of counselling delivered in assisted conception units with mild-moderate anxiety and depression delivered by skilled and experienced practitioners. This article reviews the role of counsellors as members of the assisted conception clinical team and the status of regulation and accreditation in this very new profession.

  5. Improvements in the Tl dosimetry for Radiotherapy

    International Nuclear Information System (INIS)

    The thermoluminescent dosimetry (TLD) in vivo has been demonstrated to be one of the most reliable for the control of radiotherapeutic treatments, but the delay in the response is the main disadvantage in its applicability. In this work are presented important improvements and it is demonstrated that maintaining the accuracy and reliability of the technique, it is possible to accelerate the response times at a few hours. To realize this work is utilized a lecturer Harshaw 4000, dosemeters LiF-TLD-100 chips (3.1 x 3.1 x 0.89 mm3) and rods (1 x 1 x 6 mm3). With the implementation of a glow curve analysis program developed in CIEMAT, its obtained a Tl peaks separation in such manner rapid and accurate, by that the thermal treatment of dosemeters may be reduced at one unique annealing pre-irradiation for 1 hr at 400 Centigrade. It is realized a periodical and individual calibration of the TLD and a study of the factors which influencing the ratio Tl signal-dose as linearity, correction by energy, directional response and pride of Tl signal. the results of this study are introducing in a calculation list specially designed and which allows to obtain absorbed dose by TLD starting of the dates (dosimetric peaks area) which appear of the glow curves analysis. The dose is obtained with an accuracy less than 5 %. The dosemeters already irradiated (in vivo) are analysed and informed in only four hours, allowing a greater control of the treatment and a correction of the possible errors for the next session, still in bi fractional treatments. The method implemented results thus accurate, rapid and reliable. (Author)

  6. Dosimetry audits in Brazil for {sup 192}Ir high dose rate brachytherapy systems

    Energy Technology Data Exchange (ETDEWEB)

    Rosa, L.A.R. da; Paiva, E. de.; Goncalves, M.G.; Velasco, A.F.; Di Prinzio, R.; Dovales, A.C.M.; Freire, B.L.V.; Brito, R.R.A.; Giannoni, R.A.; Castelo, L.H.R. [Instituto de Radioprotecao e Dosimetria (IRD/CNEN-RJ), Rio de Janeiro, RJ (Brazil); Marechal, M.H.H. [Comissao Nacional de Energia Nuclear (CNEN), Rio de Janeiro, RJ (Brazil). Coordenacao de Instalacoes Radiativas (CORAD)

    2005-03-15

    In Brazil, among 200 radiotherapy centres, 30 have high dose rate (HDR) {sup 192}Ir brachytherapy systems. In August 2001, the Brazilian National Nuclear Energy Commission (CNEN) started a biennial audit program to those centres having HDR systems. This program consists of visiting each centre in order to investigate the radiation protection aspects of the centres and also to measure the intensity of the brachytherapy source, in terms of air kerma strength, with a well type chamber specially designed for HDR {sup 192} Ir sources. The audit dosimetry results are compared to measurements carried out by the local institution physicist and to the source intensity value provided by the manufacturer. Two methods have been used by the Brazilian physicists for HDR {sup 192}Ir brachytherapy source dosimetry, namely the employment of a farmer type chamber calibrated according to the interpolation methodology and the use of a well type chamber to provide direct intercomparison. The larger difference obtained was 18.9% and it can be explained in terms of the lack of knowledge of the institution physicist about the interpolation methodology using the farmer type chamber. Another difference of 5.82% was found as being the lack of an updated calibration factor for the clinic well type chamber. On the basis of these results, CNEN is able to establish a maximum deviation value for the dosimetry of HDR system. Additionally, with this program the radiotherapy services have an opportunity to have their HDR {sup 192}Ir sources calibrated and to test the validity of the calibration factors for their own well type chambers, using their calibrated sources. (author)

  7. Experimental dosimetry of Ho-166 bioglass seed polymer-protected

    Energy Technology Data Exchange (ETDEWEB)

    Nogueira, Luciana B.; Campos, Tarcisio P.R. [Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil). Dept. de Engenharia Nuclear. Programa de Pos-Graduacao em Ciencias e Tecnicas Nucleares

    2011-07-01

    This study aims to develop experimental dosimetry of Ho-166 bio glass seed for brachytherapy studies using GAFCHROMIC EBT2 radio chromium films. The methodology consists of placement of radio chromium films in a compressed breast phantom, along with bio glass polymer-protected seeds of [Si: Ca: Ho] and [Si: Ca: Ho: Zr]. The bio glass seeds were encapsulated with polyvinyl alcohol, before being activated and used in the study. The bio glass seeds were introduced into the breast phantom, along with radio chromium films for a period of 2 hours. After the exposure time, radio chromium films were removed from phantom and digitized for analysis in ImageDIG 2.0 program, which quantifies the intensity of RGB (Red, Green, Blue). The dose calculation was evaluated by Monte Carlo technique. Experimental and theoretical data were used to calibrate the dose distribution. The results were plotted on graphs and dose iso curves were obtained. As conclusion it is possible to perform dosimetry in Ho-166 seed brachytherapy using radio chromium films, limited to a short exposure time and small activity. (author)

  8. Biological dosimetry of ionizing radiation in the high dose range

    International Nuclear Information System (INIS)

    The report reviews briefly methods of dose evaluation after exposure to high doses of ionizing radiation. Validation of two methods also is described: micronucleus (Mn) frequency estimation according Muller and Rode and premature chromosome condensation (PCC) combined with painting of 3 pairs of chromosomes in human lymphocytes. According to Muller and Rode, micronucleus frequency per binucleated cells with at least one Mn linearly increases with dose up to 15 Gy and is suitable end-point for biological dosimetry. These authors, however, examined cells from only one donor. The data reported below were obtained for 5 donors; they point to a considerable individual variation of thus measured response to irradiation. Due to the high degree of inter-donor variability, there is no possibility to apply this approach in biological dosimetry in the dose range 5 - 20 Gy gamma 60Co radiation. A linear response up to 10 Gy was observed only in the case of certain donors. In contrast, determination of the dose-effect relationship with the PCC method gave good results (small inter-individual variation, no plateau effect up to dose 10 Gy), so that with a calibration curve it could be used for dose estimation after exposure to doses up to 10 Gy of X or gamma 60Co radiation. (author)

  9. Internal dosimetry, past and future

    International Nuclear Information System (INIS)

    Progress in the dosimetry of internally deposited radionuclides since World War II is reviewed. The Permissible Doses Conference held at Chalk River in 1949 defined the Standard Man and a biokinetic lung model, setting maximum permissible body burdens (MPBB), maximum permissible concentrations in air (MPCA), and maximum permissible concentrations in water (MPCW) for selected radionuclides. ICRP publications 2, 6 and 9 followed, focusing on setting MPCs. The use of the power function to describe radionuclide retention in the human body was discussed in Publication 2, but not recommended for use until Publication 6. Publication 2 defined the term effective energy. The integration time for internal exposures became 50 years, and the committed dose was defined. ICRP publications 10, 10A and 54 provided guidance for the calculation of doses from measured activity in vivo or in excreta. In 1979 ICRP publication 30 replaced publication 2 as the handbook for internal dosimetry. There will be a major revision of Publication 30 following the release of the new ICRP recommendations. A future publication will give doses to patients who have been administered radiopharmaceuticals. New computer tools will allow the development of more realistic metabolic models, and new dosimetric models that calculate doses to cells will be developed. The availability of high resolution solid state detectors has resulted in improvement in measurements of radionuclides in vivo, and some improvement in radiochemical analyses of excreta. However, poor sensitivity to actinides leaves something to be desired in vitro measurements and air monitoring

  10. Dosimetry effects of film packing

    International Nuclear Information System (INIS)

    Full text: Dosimetric artefacts in film based dosimetry have been addressed by a number of authors. We have investigated the influence on film dose results, of a number of materials that are commonly packed against the film including, solid water, paper, air and plastic. The results indicate that variations in optical density occur due to the character and relative quantity of the packing material as well as the film itself. Kodak X-omat V and GAFChromic film samples were placed in a solid water cassette with packing sheets of various materials placed in contact with the film. Photon and electron exposures were carried out with various film orientation and beam qualities. Results have been obtained for solid water, paper and air. An example of the relative change in film density as a function of depth due to four paper sheets packed adjacent to a film aligned with the central axis of a 6MV photon beam is shown. Other results indicate dose variation can be attributed to Cerenkov radiation. Packing materials in contact or in close proximity with dosimetric film, contribute to optical density variations of the order of several percent. Careful consideration of these effects is necessary when using film in high accuracy dosimetry. Copyright (2001) Australasian College of Physical Scientists and Engineers in Medicine

  11. Fast neutron dosimetry: Progress summary

    International Nuclear Information System (INIS)

    The purpose was to investigate the radiological physics and biology of very low energy photons derived from a 1-GeV electron synchrotron storage ring. An extensive beam line and irradiation apparatus was designed, developed, and constructed. Dosimetry measurements required invention and testing of a miniature absolute calorimeter and a cell irradiation fixture suitable for scanning exposures under computer control. Measurements of the kerma factors of oxygen, aluminum and silicon for 14-20 MeV neutrons. Custom designed miniature proportional counters of cylindrical symmetry were employed in these determinations. The oxygen kerma factor was found significantly lower than values calculated from microscopic cross sections. We also tested Mg and Fe walled conventional spherical counters. The direct neutron-counting gas interaction is significant enough for these counters that a correction is needed. We also investigated the application of Nuclear Magnetic Resonance spectroscopy to radiation dosimetry. Our purpose was to take advantage of recent development of very high-field magnets, complex RF-pulse techniques for solvent suppression, and improved spectral analysis techniques

  12. CT dosimetry and risk estimates

    International Nuclear Information System (INIS)

    Conventional approaches to CT dosimetry are inadequate because they fail to evaluate the risk to the patient. A risk related approach to CT dosimetry is developed which explicitly takes into account the non-uniform dose distribution in the body and the relative sensitivities of different organs and tissues. The principal radiological risks to patients undergoing CT examinations are the stochastic processes of carcinogenesis and genetic effects. Radiation risk estimates have been obtained for an EMI 5005 CT scanner by measuring the mean organ doses in a Rando phantom for CT studies on the head, chest, abdomen and pelvis. The application of these risk estimates to the population served by this CT scanner indicates that during the lifetime of the CT scanner, approximately 50,000 patients will undergo CT scanning and 60% will result in a positive diagnosis. The radiation detriment is estimated to be about 1 induced cancer and a negligible genetic effect. The radiation detriment is considerably smaller than the total detriment associated with contrast material used in 76% of the CT studies. (author)

  13. Health physics research reactor reference dosimetry

    International Nuclear Information System (INIS)

    Reference neutron dosimetry is developed for the Health Physics Research Reactor (HPRR) in the new operational configuration directly above its storage pit. This operational change was physically made early in CY 1985. The new reference dosimetry considered in this document is referred to as the 1986 HPRR reference dosimetry and it replaces any and all HPRR reference documents or papers issued prior to 1986. Reference dosimetry is developed for the unshielded HPRR as well as for the reactor with each of five different shield types and configurations. The reference dosimetry is presented in terms of three different dose and six different dose equivalent reporting conventions. These reporting conventions cover most of those in current use by dosimetrists worldwide. In addition to the reference neutron dosimetry, this document contains other useful dosimetry-related data for the HPRR in its new configuration. These data include dose-distance measurements and calculations, gamma dose measurements, neutron-to-gamma ratios, ''9-to-3 inch'' ratios, threshold detector unit measurements, 56-group neutron energy spectra, sulfur fluence measurements, and details concerning HPRR shields. 26 refs., 11 figs., 31 tabs

  14. Feasibility study on introduction of KOLAS (Korea Laboratory Accreditation Scheme) in nuclear examination facility

    Energy Technology Data Exchange (ETDEWEB)

    Park, Dae Gyu; Hong, K. P.; Song, W. S.; Min, D. K

    1999-07-01

    To be an institute officially authorized by the KOLAS, the understanding and the analysis of following contents is required.: the understanding of concept required to get the accreditation of testing, the system specifying an internationally accredited testing and examination organization, international organization in the field of laboratory accreditation, domestic laboratory accreditation organization(KOLAS), the investigation of the regulations with laboratory accreditation in Korea, the investigation of the procedures accrediting a testing and examination organization, the investigation of general requirements(ISO 17025) for a testing and examination organization. (author)

  15. Biological dosimetry of X-rays by micronuclei study

    International Nuclear Information System (INIS)

    Biological dosimetry consists of estimating absorbed doses for people exposed to radiation by mean biological methods. Several indicators used are based in hematological, biochemical an cytogenetics data, although nowadays without doubt, the cytogenetic method is considered to be the most reliable, in this case, the study of micronuclei in peripheral blood lymphocytes cytokinetic blocked can be related to absorbed dose through an experimental calibration curve. An experimental dose-response curve, using micronuclei assay for X-rays at 250 kVp, 43,79 rads/min and temperature 37 degree celsius has been produced. Experimental data is fitted to model Y=c+ α D+β D2 where. Y is the number micronuclei per cell and D the dose. the curve is compared with those produced elsewhere

  16. Biological dosimetry of ionizing radiation by chromosomal aberration analysis

    International Nuclear Information System (INIS)

    Biological dosimetry consists of estimating absorbed doses for people exposed to radiation by mean biological methods. Several indicators used are based in haematological, biochemical, and cytogenetic data, although nowadays without doubt, the cytogenetic method is considered to be the most reliable. In this case, the study ol chromosomal aberrations, normally dicentric chromosomes, in peripheral lymphocytes can be related to absorbed dose through an experimental calibration curve. An experimental dose-response curve, using dicentric chromosomes analysis, X-rays at 300 kVp, 114 rad/min and temperature 37 degree celsius has been produced. Experimental data is fitted to model Y =α + β1D + β2D 2 , where Y is the number of dicentrics per cell and D the dose. The curve is compared with those produced elsewhere. (Author) 14 refs

  17. Environmental radiation monitoring around nuclear power plant using thermoluminescence dosimetry

    International Nuclear Information System (INIS)

    A field study was undertaken at the Bushehr Nuclear Power Plant, BNPP, and surrounding areas from August 1996 to August l997 as a part of a pre-operational program to establish the natural background radiation level for this region. The purpose of this project was to study environmental background radiation which was accomplished by thermoluminescence dosimetry. 18 TLD stations were set up and distributed around the nuclear power plant up to distance s of 30 km radius. The selected dosimeters consisted of two Calcium Sulfate (Titanium doped), CaSO4:Tm, disks. Dosimeters were calibrated and utilized according to the ANSI N545 -1975 standard procedure. The measured exposure rates of the background radiation level from all the stations were in the range of (0.03 - 0.12 μSv/h) with the mean value of 0.075 μSv/h. (Author)

  18. LWR-PV Surveillance Dosimetry Improvement Program review graphics

    International Nuclear Information System (INIS)

    A primary objective of the multilaboratory program is to prepare an updated and improved set of dosimetry, damage correlation, and the associated reactor analysis ASTM standards for LWR-PV irradiation surveillance programs. Supporting this objective are a series of analytical and experimental validation and calibration studies in Benchmark Neutron Fields, reactor Test Regions, and operating power reactor Surveillance Positions. These studies will establish and certify the precision and accuracy of the measurement and predictive methods which are recommended for use in these standards. Consistent and accurate measurement and data analysis techniques and methods, therefore, will have been developed and validated along with guidelines for required neutron field calculations that are used to (1) correlate changes in material properties with the characteristics of the neutron radiation field and (2) predict pressure vessel steel toughness and embrittlement from power reactor surveillance data

  19. Preparation and dosimetry of radiotherapeutic particles for arthropaties

    International Nuclear Information System (INIS)

    It was developed a new formulation of macro aggregates of Samarium 153 (153 Sm-MH) for the arthropaties treatment. The radio pharmaceutic was prepared by reaction of Samarium 153 chloride (SmCl3) in aqueous environment with sodium boron hydride in NaOH 0.5 N. The microscopic analysis shown that the particles have an average size of 4% m (range 1-14 μ m). The velocity of sedimentation was 0.008 cm/min with high stability in vitro in human serum. The biological studies in healthy rabbits, shown that the complex is retained inside the articulation still eight days after of the administration of the radiopharmaceutical. Likewise, it is presented the data of absorbed dose in the different target organs, which was determined by thermoluminescent dosimetry (TLD) through the use of a REMCAL phantom (radiation equivalent manikin calibration). (Author)

  20. Relative electron dosimetry using a synthetic diamond probe

    CERN Document Server

    Merwe, D G

    1999-01-01

    Implementation of Bragg-Gray Cavity Theory in electron dosimetry is complicated by the fact that most commercial detector volumes behave as small field inhomogeneities. Several correction factors are necessary to establish the absorbed dose at a particular point in a homogenous tissue-equivalent phantom. The energy dependence of air and the replacement effects introduced as a result of air ionization chambers' construction and size, increase the uncertainty of electron beam calibrations. Preliminary relative dose measurements performed with a prototype synthetic diamond are presented here. Theoretically, the radiation response of diamond, synthetic or natural, has negligible energy dependence. The sensitivity and small size of the probe makes it an excellent candidate for measurements in fields of high dose gradient.

  1. Biological Dosimetry of X-rays by micronuclei study

    International Nuclear Information System (INIS)

    Biological dosimetry consists of estimating absorbed doses for people exposed to radiation by mean biological methods. Several indicators used are based in haematological, biochemical an cytogenetics data, although nowadays without doubt, the cytogenetic method is considered to be the most reliable, in this case, the study of micronuclei in peripheral blood lymphocytes citokinetics blocked can be related to absorbed dose through an experimental calibration curve. An experimental dose-response curve, using micronuclei assay for X-rays at 250 kVp, 43,79 rads/min and temperature 37 degree centigree has been produced. Experimental data is fitted to model Y=C+ αD+BD''2 where Y is the number of micronuclei per cell and D the dose. The curve is compared with those produced elsewhere. (Author) 24 refs

  2. Measuring Systems for Thermometer Calibration in Low-Temperature Range

    Science.gov (United States)

    Szmyrka-Grzebyk, A.; Lipiński, L.; Manuszkiewicz, H.; Kowal, A.; Grykałowska, A.; Jancewicz, D.

    2011-12-01

    The national temperature standard for the low-temperature range between 13.8033 K and 273.16 K has been established in Poland at the Institute of Low Temperature and Structure Research (INTiBS). The standard consists of sealed cells for realization of six fixed points of the International Temperature Scale of 1990 (ITS-90) in the low-temperature range, an adiabatic cryostat and Isotech water and mercury triple-point baths, capsule standard resistance thermometers (CSPRT), and AC and DC bridges with standard resistors for thermometers resistance measurements. INTiBS calibrates CSPRTs at the low-temperature fixed points with uncertainties less than 1 mK. In lower temperature range—between 2.5 K and about 25 K — rhodium-iron (RhFe) resistance thermometers are calibrated by comparison with a standard which participated in the EURAMET.T-K1.1 comparison. INTiBS offers a calibration service for industrial platinum resistance thermometers and for digital thermometers between 77 K and 273 K. These types of thermometers may be calibrated at INTiBS also in a higher temperature range up to 550°C. The Laboratory of Temperature Standard at INTiBS acquired an accreditation from the Polish Centre for Accreditation. A management system according to EN ISO/IEC 17025:2005 was established at the Laboratory and presented on EURAMET QSM Forum.

  3. Experimental analysis of a novel and low-cost pin photodiode dosimetry system for diagnostic radiology

    Science.gov (United States)

    Nazififard, Mohammad; Suh, Kune Y.; Mahmoudieh, Afshin

    2016-07-01

    Silicon PIN photodiode has recently found broad and exciting applications in the ionizing radiation dosimetry. In this study a compact and novel dosimetry system using a commercially available PIN photodiode (BPW34) has been experimentally tested for diagnostic radiology. The system was evaluated with clinical beams routinely used for diagnostic radiology and calibrated using a secondary reference standard. Measured dose with PIN photodiode (Air Kerma) varied from 10 to 430 μGy for tube voltages from 40 to 100 kVp and tube current from 0.4 to 40 mAs. The minimum detectable organ dose was estimated to be 10 μGy with 20% uncertainty. Results showed a linear correlation between the PIN photodiode readout and dose measured with standard dosimeters spanning doses received. The present dosimetry system having advantages of suitable sensitivity with immediate readout of dose values, low cost, and portability could be used as an alternative to passive dosimetry system such as thermoluminescent dosimeter for dose measurements in diagnostic radiology.

  4. First intercomparison exercise in the frame of the coordinated investigation program of the IAEA on regional intercomparison of personal dosimetry

    International Nuclear Information System (INIS)

    During the days 7 and 11 of October of 1996 took place in Buenos Aires, Argentina, the first Meeting of the Coordinated Investigation program of the IAEA on Regional Intercomparison of Personal Dosimetry for Latin American. In this meeting participated nine representatives of reference laboratories and of personal dosimetry of the region. Fundamental aspect of personal dosimetry relates with the quantity personal dose equivalent Hp application and the implementation of intercomparison exercise in order to improve the quality of the dose estimation have been discussed. Also lectures carried out by the specialist on Hp and practical aspects of it implementation; answer and calibration according to the ISO 4037; intercomparison methods: procedures and organizations. It was carried out the first intercomparison exercise where the participants collaborated in the preparations and irradiations of personal dosemeters they have brought. (author)

  5. Intercalibration of radiological measurements for surveillance purposes of the internal dosimetry laboratory coordinated by the IAEA; Intercalibracion de mediciones radiologicas para fines de vigilancia del laboratorio de dosimetria interna coordinada por el OIEA

    Energy Technology Data Exchange (ETDEWEB)

    Alfaro L, M.M. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)

    2002-07-15

    The ININ of Mexico participated in this intercomparison organized by the IAEA in 2000. The objective of this activity is that the dosimetry laboratories that participate can validate the programs of internal dosimetry, with the purpose of improving its capacity in the evaluation of the internal dose and have access to a mechanism to evaluate its dosimetry system under real conditions. The specific objectives of this intercomparison were: 1. To evaluate the participant's capacity to manage the measurements of individual monitoring in terms of the activity in the phantom. 2. To provide the access to the unique calibration resources that otherwise would not be available. 3. To compare the operation of several detection systems, the geometry, phantoms, calibration methods and methods for the evaluation of activity of the radionuclide used by each institution. 4. To provide the independent verification of the direct measurement methods of the dosimetry service. (Author)

  6. Characterization of a 137Cs standard source for calibration purposes ar CRCN-NE

    International Nuclear Information System (INIS)

    Radiation protection monitoring instruments should be calibrated by accredited calibration laboratories. To offer calibration services, a laboratory must accomplish all requirements established by the national regulatory agency. The Calibration Service of the Centro Regional de Ciencias Nucleares (CRCN-NE), Comissao Nacional de Energia Nuclear, Recife, Brazil, is trying to achieve this accreditation. In the present work, a 137Cs standard source was characterized following the national and international recommendations and the results are presented. This source is a commercially available single source irradiator model 28-8A, manufactured by JLShepherd and Associates, with initial activity of 444 GBq (05/13/03). To provide different air kerma rates, as required for the calibration of portable radiation monitors, this irradiator have a set of four lead attenuators with different thickness, providing attenuation factors equal to 2, 4, 10 and 100 times (nominally). The performed tests included: size and uniformity of the radiation standard field at calibration reference position, variation of the air kerma rate for different lead attenuators, determination of attenuation factors for each lead attenuator configuration, and determination of the radiation scattering at the calibration reference position. The results showed the usefulness of the 137Cs standard source for the calibration of radiation protection monitoring detectors. (author)

  7. Electron paramagnetic resonance technique for radiation dosimetry: emerging trends for laboratory and accidental dosimetry

    International Nuclear Information System (INIS)

    The applications of Electron Paramagnetic Resonance (EPR) for radiation dosimetry are briefly reviewed. In particular, EPR-alanine dosimetry and accidental dosimetry using EPR signals from human tooth enamel have been discussed. The alanine dosimetry was found to be useful from low doses such as 1 Gy to high doses such as 100 kGy. The signals from tooth enamel are found to be invaluable in assessing the absorbed dose of people exposed to radiation accidents and also survivors of atomic bomb explosions. New emerging trends using EPR signals from bones exposed to radiation have also been briefly reviewed. (author)

  8. Traceable Pyrgeometer Calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina

    2016-05-02

    This poster presents the development, implementation, and operation of the Broadband Outdoor Radiometer Calibrations (BORCAL) Longwave (LW) system at the Southern Great Plains Radiometric Calibration Facility for the calibration of pyrgeometers that provide traceability to the World Infrared Standard Group.

  9. First Central and Eastern European Workshop on Quality control, patient dosimetry and radiation protection in diagnostic and interventional radiology and nuclear medicine

    International Nuclear Information System (INIS)

    First Central and Eastern European Workshop on Quality Control, Patient Dosimetry and Radiation Protection in Diagnostic and Interventional Radiology and Nuclear Medicine, scientifically supported and accredited as a CPD event for medical physicists by EFOMP, National 'Frederic Joliot-Curie' Research Institute for Radiobiology and Radiohygiene (NRIRR), Budapest, Hungary, April 25-28, 2007. Topics of the meeting included all areas of medical radiation physics except radiation therapy. A unique possibility was realized by inviting four European manufacturers of quality control instrumentation, not only for exhibiting but they also had 45 minutes individual presentations about each manufacturer's product scale and conception. Further sessions dealt with dosimetry, optimization, quality control and testing, radiation protection and standardization, computed tomography and nuclear medicine, in 29 oral presentations and 1 poster of the participants. (S.I.)

  10. On the calibration of radiotherapy dosemeters in Australia

    International Nuclear Information System (INIS)

    Full text: Dosemeters for external beam radiotherapy are calibrated in Australia by ARPANSA, against the national primary standards of exposure and absorbed dose. The primary standards are free air chambers for exposure at low and medium energy X-rays, a graphite cavity chamber for exposure at 60Co, and a graphite calorimeter for absorbed dose at 60Co and high energy (MV) X -rays. Radiotherapy dosemeters are calibrated against these standards using a well documented formalism to provide calibration factors suitable for use with dosimetry protocols. A dosemeter usually comprises an ionization chamber connected to an independent electrometer. These are calibrated separately if possible. A combined calibration factor is reported together with the electrometer calibration factor (sensitivity). The dosimetry protocol used in radiotherapy centres in Australia and New Zealand is currently the simplified version of the IAEA TRS277 protocol, published by the New Zealand NRL and recommended by the ACPSEM. This protocol requires the use of an exposure or air kerma calibration factor at 60Co (Nx or Nk) to evaluate the absorbed dose to air calibration factor ND. The chamber is then placed in a water phantom with its centre displaced from the reference point by peff. ARPANSA can also supply calibration factors in absorbed dose to water (ND,w), as required as input to the new IAEA CoP. If an absorbed dose to water calibration factor is used by the radiotherapy centre, the chamber should be placed with its centre at the reference point in the water phantom. ARPANSA has for some years coordinated the participation of Australian radiotherapy centres in the IAEA TLD Quality Audit service. Note that this service does not represent a calibration and should not be referred to as such. The only calibration is that provided by ARPANSA for a reference dosemeter at each radiotherapy centre. As soon as the ANSTO SSDL is operational, calibrations of reference dosemeters will also be available

  11. Dosimetry investigation of MOSFET for clinical IMRT dose verification.

    Science.gov (United States)

    Deshpande, Sudesh; Kumar, Rajesh; Ghadi, Yogesh; Neharu, R M; Kannan, V

    2013-06-01

    In IMRT, patient-specific dose verification is followed regularly at each centre. Simple and efficient dosimetry techniques play a very important role in routine clinical dosimetry QA. The MOSFET dosimeter offers several advantages over the conventional dosimeters such as its small detector size, immediate readout, immediate reuse, multiple point dose measurements. To use the MOSFET as routine clinical dosimetry system for pre-treatment dose verification in IMRT, a comprehensive set of experiments has been conducted, to investigate its linearity, reproducibility, dose rate effect and angular dependence for 6 MV x-ray beam. The MOSFETs shows a linear response with linearity coefficient of 0.992 for a dose range of 35 cGy to 427 cGy. The reproducibility of the MOSFET was measured by irradiating the MOSFET for ten consecutive irradiations in the dose range of 35 cGy to 427 cGy. The measured reproducibility of MOSFET was found to be within 4% up to 70 cGy and within 1.4% above 70 cGy. The dose rate effect on the MOSFET was investigated in the dose rate range 100 MU/min to 600 MU/min. The response of the MOSFET varies from -1.7% to 2.1%. The angular responses of the MOSFETs were measured at 10 degrees intervals from 90 to 270 degrees in an anticlockwise direction and normalized at gantry angle zero and it was found to be in the range of 0.98 ± 0.014 to 1.01 ± 0.014. The MOSFETs were calibrated in a phantom which was later used for IMRT verification. The measured calibration coefficients were found to be 1 mV/cGy and 2.995 mV/cGy in standard and high sensitivity mode respectively. The MOSFETs were used for pre-treatment dose verification in IMRT. Nine dosimeters were used for each patient to measure the dose in different plane. The average variation between calculated and measured dose at any location was within 3%. Dose verification using MOSFET and IMRT phantom was found to quick and efficient and well suited for a busy radiotherapy

  12. Twelfth Symposium on Neutron and Ion Dosimetry - NEUDOS-12. Editorial

    International Nuclear Information System (INIS)

    ), and a workshop on Biologically Weighted Quantities in Radiotherapy (European project BioQuaRT), chaired by Hans Rabus from German National Metrology Institute (PTB). A best poster Award was presented to Michal Sadel, from the Institute of Nuclear Physics, of the Polish Academy of Science (Poland), for his poster dealing with 'Relative TL and OSL efficiency to protons of various dosimetric materials'. The scientific programme of this 12. NEUDOS symposium covered a wide variety of topics related to neutron and ion dosimetry, which were organised into six sessions: I-Basic aspects and new issues, II-Instrumentation and techniques, III-Neutrons and ions in medicine, IV-Neutrons and complex radiation fields in research and industry, V-Complex radiation fields in aviation and space activities and VI-Physics-biology interface. It is interesting to note that >50 % of the posters were related to the topics of Session II (Dosimetry, Metrology and Spectrometry, Calibrations and Facilities, and Intercomparisons). A total of 134 manuscripts were submitted for publication in these proceedings. (authors)

  13. Alanine - ESR dosimetry, feasibility and possible applications

    International Nuclear Information System (INIS)

    Alanine ESR dosimetry presents a great interest for quality controls in radiotherapy. This new developed water equivalent alanine dosimeter allows a reproducible dose measurement, by a non-destructive readout technique in a large dose range. In this paper the stability of the dosimeter response has been shown but also its independence with the energy or the dose rate of the absorbed radiation. Through this different studies, one can broaden the application field of alanine / ESR dosimetry especially for in-vivo dosimetry. The results of the experiments and the intra operative treatment, indicate that this kind of dosimetry seems to be a promising technique for in-vivo quality controls in electron beam, γ ray or X ray radiotherapy. (authors)

  14. Photovoltaic module certification/laboratory accreditation criteria development: Implementation handbook

    Energy Technology Data Exchange (ETDEWEB)

    Osterwald, C.R. [National Renewable Energy Laboratory, Golden, CO (United States); Hammond, R.L.; Wood, B.D.; Backus, C.E.; Sears, R.L. [Arizona State Univ., Tempe, AZ (United States); Zerlaut, G.A. [SC-International, Inc., Tempe, AZ (United States); D`Aiello, R.V. [RD Associates, Tempe, AZ (United States)

    1996-08-01

    This document covers the second phase of a two-part program. Phase I provided an overview of the structure and function of typical product certification/laboratory accreditation programs. This report (Phase H) provides most of the draft documents that will be necessary for the implementation of a photovoltaic (PV) module certification/laboratory accreditation program. These include organizational documents such as articles of incorporation, bylaws, and rules of procedure, as well as marketing and educational program documents. In Phase I, a 30-member criteria development committee was established to guide, review and reach a majority consensus regarding criteria for a PV certification/laboratory accreditation program. Committee members represented PV manufacturers, end users, standards and codes organizations, and testing laboratories. A similar committee was established for Phase II; the criteria implementation committee consisted of 29 members. Twenty-one of the Phase I committee members also served on the Phase II committee, which helped to provide program continuity during Phase II.

  15. Accreditation of testing laboratories in CNEA (National Atomic Energy Commission)

    International Nuclear Information System (INIS)

    The recognition of the technical capability of a testing laboratory is carried out by Laboratory Accreditation Bodies as the result of a satisfactory evaluation and the systematic follow up of the certified qualification. In Argentina the creation of a National Center for the Accreditation of Testing Laboratories, as a first step to assess a National Accreditation System is currently projected. CNEA, as an institution involved in technological projects and in the development and production of goods and services, has adopted since a long time ago quality assurance criteria. One of their requirements is the qualification of laboratories. Due to the lack of a national system, a Committee for the Qualification of Laboratories was created jointly by the Research and Development and Nuclear Fuel Cycle Areas with the responsibility of planning and management of the system evaluation and the certification of the quality of laboratories. The experience in the above mentioned topics is described in this paper. (author)

  16. System Quality Management in Software Testing Laboratory that Chooses Accreditation

    Directory of Open Access Journals (Sweden)

    Yanet Brito R.

    2013-12-01

    Full Text Available The evaluation of software products will reach full maturity when executed by the scheme and provides third party certification. For the validity of the certification, the independent laboratory must be accredited for that function, using internationally recognized standards. This brings with it a challenge for the Industrial Laboratory Testing Software (LIPS, responsible for testing the products developed in Cuban Software Industry, define strategies that will permit it to offer services with a high level of quality. Therefore it is necessary to establish a system of quality management according to NC-ISO/IEC 17025: 2006 to continuously improve the operational capacity and technical competence of the laboratory, with a view to future accreditation of tests performed. This article discusses the process defined in the LIPS for the implementation of a Management System of Quality, from the current standards and trends, as a necessary step to opt for the accreditation of the tests performed.

  17. Advanced materials in radiation dosimetry

    CERN Document Server

    Bruzzi, M; Nava, F; Pini, S; Russo, S

    2002-01-01

    High band-gap semiconductor materials can represent good alternatives to silicon in relative dosimetry. Schottky diodes made with epitaxial n-type 4 H SiC and Chemical Vapor Deposited diamond films with ohmic contacts have been exposed to a sup 6 sup 0 Co gamma-source, 20 MeV electrons and 6 MV X photons from a linear accelerator to test the current response in on-line configuration in the dose range 0.1-10 Gy. The released charge as a function of the dose and the radiation-induced current as a function of the dose-rate are found to be linear. No priming effects have been observed using epitaxial SiC, due to the low density of lattice defects present in this material.

  18. Development of radiation biological dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Cho, Chul Koo; Kim, Tae Hwan; Lee, Yun Sil; Son, Young Sook; Kim, Soo Kwan; Jang, Won Suk; Le, Sun Joo; Jee, Young Heun; Jung, Woo Jung

    1999-04-01

    Up until now, only a few methods have been developed for radiation biological dosimetry such as conventional chromosome aberration and micronucleus in peripheral blood cell. However, because these methods not only can be estimated by the expert, but also have a little limitation due to need high technique and many times in the case of radiation accident, it is very difficult to evaluate the absorbed dose of victims. Therefore, we should develop effective, easy, simple and rapid biodosimetry and its guideline (triage) to be able to be treated the victims as fast as possible. We established the premature chromosome condensation assay and apoptotic fragment assay which was the significant relationship between dose and cell damages to evaluate the irradiation dose as correct and rapid as possible using lymphocytes and crypt cells, and compared with conventional chromosome aberration assay and micronuclei assay.

  19. Development of radiation biological dosimetry

    International Nuclear Information System (INIS)

    Up until now, only a few methods have been developed for radiation biological dosimetry such as conventional chromosome aberration and micronucleus in peripheral blood cell. However, because these methods not only can be estimated by the expert, but also have a little limitation due to need high technique and many times in the case of radiation accident, it is very difficult to evaluate the absorbed dose of victims. Therefore, we should develop effective, easy, simple and rapid biodosimetry and its guideline (triage) to be able to be treated the victims as fast as possible. We established the premature chromosome condensation assay and apoptotic fragment assay which was the significant relationship between dose and cell damages to evaluate the irradiation dose as correct and rapid as possible using lymphocytes and crypt cells, and compared with conventional chromosome aberration assay and micronuclei assay

  20. Audits for advanced treatment dosimetry

    International Nuclear Information System (INIS)

    Radiation therapy has advanced rapidly over the last few decades, progressing from 3D conformal treatment to image-guided intensity modulated therapy of several different flavors, both 3D and 4D and to adaptive radiotherapy. The use of intensity modulation has increased the complexity of quality assurance and essentially eliminated the physicist's ability to judge the validity of a treatment plan, even approximately, on the basis of appearance and experience. Instead, complex QA devices and procedures are required at the institutional level. Similarly, the assessment of treatment quality through remote and on-site audits also requires greater sophistication. The introduction of 3D and 4D dosimetry into external audit systems must follow, to enable quality assurance systems to perform meaningful and thorough audits

  1. 77 FR 51540 - Medicare Program; Approved Renewal of Deeming Authority of the Accreditation Association for...

    Science.gov (United States)

    2012-08-24

    ... Authority of the Accreditation Association for Ambulatory Health Care, Inc. for Medicare Advantage Health... Medicare Advantage ``deeming authority'' of the Accreditation Association for Ambulatory Health Care, Inc... the Medicare program, eligible beneficiaries may receive covered services through a Medicare...

  2. NATIONAL ENVIRONMENTAL LABORATORY ACCREDITATION CONFERENCE; CONSTITUTION, BYLAWS AND STANDARDS: APPROVED MAY 25, 2001

    Science.gov (United States)

    The principles and operating procedures for the National Environmental Laboratory Accreditation Conference (NELAC) are contained in the NELAC Constitution and Bylaws. The major portion of this document (standards) contains detailed requirements for accrediting environmental labo...

  3. 75 FR 51464 - Medicare and Medicaid Programs; Approval of the American Association for Accreditation of...

    Science.gov (United States)

    2010-08-20

    ... American Association for Accreditation of Ambulatory Surgery Facilities for Continued Deeming Authority for Ambulatory Surgical Centers AGENCY: Centers for Medicare & Medicaid Services (CMS). ACTION: Final notice... for Accreditation of Ambulatory Surgery Facilities' (AAAASF) request for continued recognition as...

  4. 76 FR 22709 - Medicare and Medicaid Programs; Approval of the American Association for Accreditation of...

    Science.gov (United States)

    2011-04-22

    ... HUMAN SERVICES Centers for Medicare & Medicaid Services Medicare and Medicaid Programs; Approval of the American Association for Accreditation of Ambulatory Surgery Facilities, Inc. for Deeming Authority for... decision to approve the American Association for Accreditation of Ambulatory Surgery Facilities...

  5. 77 FR 19290 - Medicare and Medicaid Programs; Renewal of Deeming Authority of the Accreditation Association for...

    Science.gov (United States)

    2012-03-30

    ... Deeming Authority of the Accreditation Association for Ambulatory Health Care, Inc. for Medicare Advantage... the Medicare Advantage ``deeming authority'' of the Accreditation Association for Ambulatory Health... covered services through a Medicare Advantage (MA) organization that contracts with CMS. The...

  6. Occupational exposure to the whole body, extremities and to the eye lens in interventional radiology in Poland, as based on personnel dosimetry records at IFJ PAN

    Science.gov (United States)

    Szumska, Agnieszka; Budzanowski, M.; Kopeć, R.

    2014-11-01

    We report results of measurements of Hp(10) from whole body dosimeters (about 53 thousand readouts), of Hp(0.07) from finger ring dosimeters (23 thousand readouts) and of Hp(3) from eye lens dosimeters (100 readouts), issued in the years 2010-12 to over 150 medical departments in Poland which apply X-rays in radiology, interventional radiology (haemodynamic, angiology, cardiac surgery), urology, orthopaedics, electrophysiology or electro-cardiology. In all measurements thermoluminescence detectors (TLD) were used: the well-known standard MTS-N (LiF:Mg, Ti) for whole body and extremity dosimetry, and the high-sensitivity MCP-N (LiF:Mg, Cu, P) for eye lens dosimetry and environmental monitoring. We analysed the data base of the accredited Laboratory of Individual and Environmental Dosimetry (LADIS) at the Institute of Nuclear Physics PAN which offers its dosimetry service to these departments on a regular basis. We found that in the population of radiation workers that studied over the years 2010-2012 in 84%, 87%, and 34% of Hp(10), Hp(0.07) and Hp(3) measurements, respectively, the level of 0.1 mSv/quarter did not exceed, indicating lack of their occupational exposure. In the remaining 16%, 13% and 66% of individual cases, the 0.1 mSv/quarter exceeded, occasionally reaching several hundreds of mSv/quarter.

  7. The international protocol for the dosimetry of external radiotherapy beams based on standards of absorbed dose to water

    International Nuclear Information System (INIS)

    An International Code of Practice (CoP, or dosimetry protocol) for external beam radiotherapy dosimetry based on standards of absorbed dose to water has been published by the IAEA on behalf of IAEA, WHO, PAHO and ESTRO. The CoP provides a systematic and internationally unified approach for the determination of the absorbed dose to water in reference conditions with radiotherapy beams. The development of absorbed-dose-to-water standards for high-energy photons and electrons offers the possibility of reducing the uncertainty in the dosimetry of radiotherapy beams. Many laboratories already provide calibrations at the radiation quality of 60Co gamma-rays and some have extended calibrations to high-energy photon and electron beams. The dosimetry of kilovoltage x-rays, as well as that of proton and ion beams can also be based on these standards. Thus, a coherent dosimetry system based on the same formalism is achieved for practically all radiotherapy beams. The practical use of the CoP as simple. The document is formed by a set of different CoPs for each radiation type, which include detailed procedures and worksheets. All CoPs are based on ND,w chamber calibrations at a reference beam quality Qo, together with radiation beam quality correction factors kQ preferably measured directly for the user's chamber in a standards laboratory. Calculated values of kQ are provided together with their uncertainty estimates. Beam quality specifiers are 60Co, TPR20,10 (high-energy photons), R50 (electrons), HVL and kV (x-rays) and Rres (protons and ions)

  8. Application of radiation damage effects in dosimetry

    International Nuclear Information System (INIS)

    some general aspects of radiation dosimetry are outlined. The techniques of radiophotoluminescence, radiothermoluminescence and exo-electron emission are discussed individually. It is thought that the trend in personnel dosimetry is such that thermoluminescence will steadily replace film and photoluminescence techniques over the next decade, and that more unusual techniques, such as exo-electron emission, will make inroads only for special purposes. (B.R.H.)

  9. Developments in Patient Dosimetry for Unsealed Sources

    International Nuclear Information System (INIS)

    In molecular radiotherapy, treatment planning essentially is the determination of the activity to administer to optimize safety and efficacy of a treatment. Individualization is possible, for example, by using quantitative imaging modalities, external counting and blood sampling for pre-therapeutic biokinetics measurements. Patient specific dosimetry can be performed as in radiation therapy. Over- or under treatment of patients can be avoided. Here, the standard methods and the expected advances in performing individualized dosimetry are discussed. (author)

  10. Radiochromic film calibration wedge EBT2 using virtual fields; Calibracion de peliculas radiocromicos EBT2 mediante campos con cunas virtuales

    Energy Technology Data Exchange (ETDEWEB)

    Perez, M. A.; Macias, J.; Merchan, M. A.; Campo, J. L.; Moreno, J. C.; Terron, J. A.; Miras, H.; Ortiz, M.; Arrans, R.; Ortiz, A.; Fernandez, D.

    2011-07-01

    EBT2 film dosimetry after exposure to a gradient of these wedge dosimetry. In our case a virtual wedge 600. The primary objective is to automate the process, reduce the time spent in obtaining the calibration curve (color-dose). Time negligible due to the limited availability of accelerators. This method of obtaining the calibration curve provides similar results to the commonly accepted either with irradiation uniform of a single film with different dose levels (multiband ladder) or with irradiation uniform of small rectangular piece of film , decreasing by a factor about 20, the time spent. (Author)

  11. A new implementation of multichannel radiochromic film dosimetry; Una nueva implementacion del analisis multicanal para la dosimetria mediante peliculas radiocromicas

    Energy Technology Data Exchange (ETDEWEB)

    Rodriguez Rodriguez, C.; Martin Martin, G.; Bermudez Luna, R.; Lopez Fernandez, A.; Tores Olombrada, M. V. de

    2014-07-01

    The aims of this paper are to carry out a new implementation of the multichannel radiochromic film dosimetry (Micke A, Lewis D, Yu X. Multichannel film dosimetry with nonuniformity correction. Med Phys 2011;38:2523-34), to quantify the variation in gamma index as compared to the single channel film dosimetry, and to determine if the procedure achieves similar results by means of a different scanner that the one used by Micke et al. Radiochromic EBT2 films and a Microtek 9000 XL scanner were used. Our procedure simplifies the system calibration splitting it into two factors, manufactured batch and digitalization specific. Absorbed dose spatial distributions from an open radiotherapy beam without any modulation and 20 IMRT treatments were determined. Their gamma index maps were calculated and a comparison of the results from single channel and multichannel dosimetry was performed. A 5% mean increase in concordance was obtained by using the multichannel film dosimetry. Our results are similar to those reported by Micke et al. even though we are using a different scanner. (Author)

  12. Injuries and Safe Communities Accreditation: Is there a link?

    Science.gov (United States)

    Sinelnikov, Sergey; Friedman, Lee S; Chavez, Emily A

    2016-06-01

    Safe Communities (SC) is a global movement that brings together community stakeholders to collaboratively address injury concerns. SC accreditation is a formal process through which communities are recognized for strengthening local injury prevention capacity. Six million Americans live in 25 SC sites, but no research has been done to understand the model's potential impact on this population. This study explored the temporal relationship between SC accreditation and injury trends in three SC sites from the state of Illinois-Arlington Heights, Itasca, and New Lenox. Hospitalization data, including patient demographics, exposure information, injury outcomes, and economic variables, were obtained from a statewide hospital discharge database for a 12-year period (1999-2011). Joinpoint regression models were fitted to identify any periods of significant change, examine the direction of the injury trend, and to estimate monthly percent changes in injury counts and rates. Poisson random-intercept regression measured the average total change since the official SC accreditation for the three communities combined and compared them to three matched control sites. In joinpoint regression, one of the SC sites showed a 10-year increase in hospitalization cases and rates followed by a two-year decline, and the trend reversal occurred while the community was pursuing the SC accreditation. Injury hospitalizations decreased after accreditation compared to the pre-accreditation period when SC sites were compared to their control counterparts using Poisson modeling. Our findings suggest that the SC model may be a promising approach to reduce injuries. Further research is warranted to replicate these findings in other communities.

  13. Thermoluminescence dosimetry quality assurance network for radiotherapy and radiology in the Czech Republic

    International Nuclear Information System (INIS)

    The Czech thermoluminescence dosimetry (TLD) quality assurance network provides independent audits for external beam radiotherapy and dental X ray radiology, primarily for the purposes of State supervision. Two modes of audit are used in radiotherapy. The basic mode includes beam calibration checks for all clinically used photon and electron beams, while the advanced mode covers TLD measurements for photon beams both on and off the central beam axis for a few simple treatment set-ups using a multipurpose phantom. The audit aims to check for dental radiology not only some basic dosimetric characteristics of the X ray apparatus but also the film developing process used by the dentists.The audits are performed by means of mailed dosimetry. (author)

  14. High-dose dosimetry of beta rays using blue beryl dosimeters

    Energy Technology Data Exchange (ETDEWEB)

    Carmo, Lucas S. do, E-mail: lsatiro@usp.br [Instituto de Pesquisas Energeticas e Nucleares (IPEN/CNEN-SP), Sao Paulo, SP (Brazil); Watanabe, Shigueo; Bittencour, Jose F., E-mail: Lacifid@if.usp.br [Universidade de Sao Paulo (USP), Sao Paulo, SP (Brazil). Instituto de Fisica. Departamento de Fisica Nuclear

    2015-07-01

    High dose radiation is widely used in industrial applications as sterilization of medical products, improvement of materials properties, color enhancement of jewelry stones, etc. The radiation dosimetry of high doses is quite important for these applications. In this work we have investigated the usage of blue beryl crystal also known as aquamarine in high dose dosimetry of beta rays. Some works have shown that silicate minerals exhibit a good Thermoluminescent response when irradiated up to 2000 kGy of gamma rays. Here, we have produced small beryl pellets of approximately 5 mm in diameter and 3 mm thickness to measure high doses of beta rays produced at an electron accelerator at IPEN. Twelve beryl dosimeters were made and six of them were irradiated from 10kGy up to 100 kGy. The technique used to create a calibration curve was the thermoluminescence using the glow peak at 310°C. (author)

  15. 78 FR 59701 - Medicare Program; Approval of Accrediting Organization for Suppliers of Advanced Diagnostic...

    Science.gov (United States)

    2013-09-27

    ... organization's knowledge and experience in the advanced diagnostic imaging arena. A plan for reducing the... Organization for Suppliers of Advanced Diagnostic Imaging Supplier Accreditation Program AGENCY: Centers for...Site TM , a national accreditation organization to accredit suppliers seeking to furnish the...

  16. A Pillar for Successful Business School Accreditation: Conducting the Curriculum Review Process--A Systematic Approach

    Science.gov (United States)

    Gundersen, David E.; Jennings, Susan Evans; Dunn, Deborah; Fisher, Warren; Kouliavtsev, Mikhail; Rogers, Violet

    2011-01-01

    The Association to Advance Collegiate Schools of Business (AACSB) describes their accreditation as the "hallmark of business education." According to information at BestBizSchools.com (n.d.), "AACSB accreditation represents the highest standard of achievement for business schools worldwide. Being AACSB accredited means a business…

  17. Managing the Demands of Accreditation: The Impact on Global Business Schools

    Science.gov (United States)

    Kourik, Janet L.; Maher, Peter E.; Akande, Benjamin O.

    2011-01-01

    Over the past several years the academic community has become abundantly aware of the requirements of university-wide and specialized accreditation. This paper describes the background to accreditation models initiated in several regions of the world, such as the specialized business accreditations of the European Quality Improvement System…

  18. 78 FR 56711 - Health Insurance Exchanges; Application by the Accreditation Association for Ambulatory Health...

    Science.gov (United States)

    2013-09-13

    ... accreditation program to conduct surveys for ambulatory surgery centers that wish to participate in the Medicare... HUMAN SERVICES Centers for Medicare & Medicaid Services Health Insurance Exchanges; Application by the Accreditation Association for Ambulatory Health Care To Be a Recognized Accrediting Entity for the...

  19. Role of accrediting bodies in providing education leadership in medical education

    Directory of Open Access Journals (Sweden)

    Sam Leinster

    2014-01-01

    Role of accreditation authorities: If accreditation authorities are to provide leadership in medical education they must undertake regular review of their standards. This should be informed by all stakeholders and include experts in medical education. The format of the standards must provide clear direction to medical schools. Accreditation should take place regularly and should result in the production of a publicly accessible report.

  20. 78 FR 9899 - National Committee on Foreign Medical Education and Accreditation

    Science.gov (United States)

    2013-02-12

    ... National Committee on Foreign Medical Education and Accreditation AGENCY: Office of Postsecondary Education, U.S. Department of Education, National Committee on Foreign Medical Education and Accreditation... Foreign Medical Education and Accreditation (NCFMEA). Parts of this meeting will be open to the...

  1. 78 FR 45917 - National Committee on Foreign Medical Education and Accreditation Meeting

    Science.gov (United States)

    2013-07-30

    ... National Committee on Foreign Medical Education and Accreditation Meeting AGENCY: Office of Postsecondary Education, National Committee on Foreign Medical Education and Accreditation, U.S. Department of Education... Foreign Medical Education and Accreditation (NCFMEA). Parts of this meeting will be open to the...

  2. 77 FR 13312 - National Committee on Foreign Medical Education and Accreditation

    Science.gov (United States)

    2012-03-06

    ... National Committee on Foreign Medical Education and Accreditation AGENCY: National Committee on Foreign Medical Education and Accreditation, Office of Postsecondary Education, U.S. Department of Education... Foreign Medical Education and Accreditation (NCFMEA). Parts of this meeting will be open to the...

  3. 77 FR 49788 - National Committee on Foreign Medical Education and Accreditation

    Science.gov (United States)

    2012-08-17

    ... National Committee on Foreign Medical Education and Accreditation AGENCY: Office of Postsecondary Education, U.S. Department of Education, National Committee on Foreign Medical Education and Accreditation... Foreign Medical Education and Accreditation (NCFMEA). Parts of this meeting will be open to the...

  4. Op weg naar een accreditatiesysteem van Nederlandse ziekenhuizen [Towards an accreditation system of Dutch hospitals

    NARCIS (Netherlands)

    Gennip, E.M.S.J. van; Linnebank, F.; Sillevis Smitt, P.A.E.; Geldof, C.A.

    1999-01-01

    The development of the Netherlands system for accreditation of hospitals started in 1989 in the Pilotproject Accreditation (PACE). This resulted in the establishment of the Netherlands Institute for Accreditation of Hospitals (NIAH) early 1999, by the Dutch Association of Hospitals, the Dutch Associ

  5. 42 CFR 493.573 - Continuing Federal oversight of private nonprofit accreditation organizations and approved State...

    Science.gov (United States)

    2010-10-01

    ... CERTIFICATION LABORATORY REQUIREMENTS Accreditation by a Private, Nonprofit Accreditation Organization or... nonprofit accreditation organizations and approved State licensure programs. (a) Comparability review. In... 42 Public Health 5 2010-10-01 2010-10-01 false Continuing Federal oversight of private...

  6. 7 CFR 205.640 - Fees and other charges for accreditation.

    Science.gov (United States)

    2010-01-01

    ... MARKETING SERVICE (Standards, Inspections, Marketing Practices), DEPARTMENT OF AGRICULTURE (CONTINUED... other charges for accreditation. Fees and other charges equal as nearly as may be to the cost of the accreditation services rendered under the regulations, including initial accreditation, review of annual...

  7. Micrometer-resolved film dosimetry using a microscope in microbeam radiation therapy

    Energy Technology Data Exchange (ETDEWEB)

    Bartzsch, Stefan, E-mail: stefan.bartzsch@icr.ac.uk; Oelfke, Uwe [The Institute of Cancer Research, 15 Cotswold Road, Sutton SM2 5NG (United Kingdom); Lott, Johanna; Welsch, Katrin [Deutsches Krebsforschungszentrum, Im Neuenheimer Feld 280, D-69120 Heidelberg (Germany); Bräuer-Krisch, Elke [European Synchrotron Radiation Facility, 6 rue Jules Horowitz, Grenoble Cedex 9 38043 (France)

    2015-07-15

    Purpose: Microbeam radiation therapy (MRT) is a still preclinical tumor therapy approach that uses arrays of a few tens of micrometer wide parallel beams separated by a few 100 μm. The production, measurement, and planning of such radiation fields are a challenge up to now. Here, the authors investigate the feasibility of radiochromic film dosimetry in combination with a microscopic readout as a tool to validate peak and valley doses in MRT, which is an important requirement for a future clinical application of the therapy. Methods: Gafchromic{sup ®} HD-810 and HD-V2 films are exposed to MRT fields at the biomedical beamline ID17 of the European Synchrotron Radiation Facility (ESRF) and are afterward scanned with a microscope. The measured dose is compared with Monte Carlo calculations. Image analysis tools and film handling protocols are developed that allow accurate and reproducible dosimetry. The performance of HD-810 and HD-V2 films is compared and a detailed analysis of the resolution, noise, and energy dependence is carried out. Measurement uncertainties are identified and analyzed. Results: The dose was measured with a resolution of 5 × 1000 μm{sup 2} and an accuracy of 5% in the peak and between 10% and 15% in the valley region. As main causes for dosimetry uncertainties, statistical noise, film inhomogeneities, and calibration errors were identified. Calibration errors strongly increase at low doses and exceeded 3% for doses below 50 and 70 Gy for HD-V2 and HD-810 films, respectively. While the grain size of both film types is approximately 2 μm, the statistical noise in HD-V2 is much higher than in HD-810 films. However, HD-810 films show a higher energy dependence at low photon energies. Conclusions: Both film types are appropriate for dosimetry in MRT and the microscope is superior to the microdensitometer used before at the ESRF with respect to resolution and reproducibility. However, a very careful analysis of the image data is required

  8. Activities on calibration of radiation protection instruments in Indonesia

    International Nuclear Information System (INIS)

    As the use of the ionizing radiation emitted by radionuclides or produced by modern machines in Indonesia has increased significantly in the past two decades, the demand for radiation protection measures has also grown up very rapidly. In the mind of Indonesian people, ionizing radiation is always associated with atomic bombs. Indonesian government has set up National Atomic Energy Agency (BATAN) through the Act No. 31/1964. The BATAN has responsibility in the research and development, implementation and inspection of the safe use of ionizing radiation for peaceful purposes, and always put a great concern on radiation protection matter. The Center for Standardization and Radiation Safety Research (CSRSR) has been founded to implement research and services in the fields of radiation safety, standardization, dosimetry, radiation health, as well as the application of nuclear techniques to medicine. In order to provide the national reference in terms of radiation dosimetry and calibration, the Secondary Standard Dosimetry Laboratory was completely set up in Jakarta by 1984. As available facilities, radiation instruments and radiation sources are described. Calibration and personal monitoring services are reported. (K.I.)

  9. Calibration of high-dose radiation facilities (Handbook)

    International Nuclear Information System (INIS)

    In India at present several high intensity radiation sources are used. There are 135 teletheraphy machines and 65 high intensity cobalt-60 sources in the form of gamma chambers (2.5 Ci) and PANBIT (50 Ci). Several food irradiation facilities and a medical sterilization plant ISOMED are also in operation. The application of these high intensity sources involve a wide variation of dose from 10 Gy to 100 kGy. Accurate and reproducible radiation dosimetry is essential in the use of these sources. This handbook is especially compiled for calibration of high-dose radiation facilities. The first few chapters discuss such topics as interaction of radiation with matter, radiation chemistry, radiation processing, commonly used high intensity radiation sources and their special features, radiation units and dosimetry principles. In the chapters which follow, chemical dosimeters are discussed in detail. This discussion covers Fricke dosimeter, FBX dosimeter, ceric sulphate dosimeter, free radical dosimetry, coloured indicators for irrdiation verification. A final chapter is devoted to practical hints to be followed in calibration work. (author)

  10. SEVENTH DOE WORKSHOP ON PERSONNEL NEUTRON DOSIMETRY

    Energy Technology Data Exchange (ETDEWEB)

    Vallario, E J

    1978-10-24

    This workshop was the seventh of a series and was held on October 23-24. 1978, at the Central Electricity Generating Board, HQ, London, England. Typically~ attendees at the Workshop were concerned with one of three activities: studying and refining existing techniques in an attempt to quantify already-known parameters with greater precision, looking for ways to apply existing neutron dosirr:etry techniques to a specific local problem, identifying the needs and weaknesses of existing systems, with the goal of improving and passibly simplifying field measurements. The types of neutron dosimetry techniques discussed by participants included albedo dosimeters, track etch, and TLD. One speaker reported on NTA film, noting that fading could be eliminated by drying the emulsion in dry nitrogen before field use. There were no reports on tissue equivalent proportional counters or activation analysis. One participant discussed a metal oxide silicon dosimeter. The need to develop a consistent standard terminology, as well as calibration sources and techniques, on both the national and international level was evident. The need for standardization is particularly acute in the U.S. Techniques for evaluating dosimeter response in the field should he standardized, since several different instruments with widely different response characteristics are currently being used. The choice of instruments is often parochial. Also. the type and use of phantoms should be standardized. Neutron dose assignment is significantly affected by the position of the dosimeter on the body. for example, a typical albedo dosimeter may give differences of up to 20% depending on whether it is worn on the belt or chest. Larger errors are encountered with front-to-back (angular} orientation. 1n an attempt to minimize such errors~ at least two European facilities are using neutron dosimeter belts, which provide dosimeters both in front and in back of the wearer. The gamma-to-neutron ratio around nuclear power

  11. Imaging based, patient specific dosimetry

    International Nuclear Information System (INIS)

    Full text: The prognosis of achieving longtime remission for disseminated cancer disease is in many cases poor. A systemic treatment is required and therefore external beam radiation therapy is less suited. Treatment with radiolabeled pharmaceuticals, so called radionuclide therapy is such a systemic treatment. In radionuclide therapy, the absorbed dose is delivered by administration of radionuclides that emit electrons or alpha particles. It is here assumed that the released kinetic energy is transferred by interactions to sensitive parts of the cells activating cell death, and thus an accurate dosimetry is important. However, absorbed dose planning for radionuclide therapy is a real challenge in that the source cannot be turned on or off (as in external beam therapy) but decays exponentially with characteristics depending on the biokinetics and the radionuclide half-life. On a small-scale, the radiopharmaceutical is also heterogeneously distributed which means that the energy deposition is generally nonuniform. The biokinetics may also change over time which means that activity measurements need to be made at several time points to estimate the total amount of released energy in an organ or tumour. Practical issues regarding the number of measurements and patient mobility may therefore limit the accuracy in this calculation. The dose-rate for radionuclide therapy is also much lower than in external beam therapy. Since the treatment is systemic, circulating activity may result in absorbed doses to normal organs and tissues. Often this poses a problem and puts a limit on the amount of activity to can be administered. This is one of the major reasons for the requirement of an accurate patient-specific dosimetry. One of the major problems is that the biokinetics varies between patients and the activity uptake and clearance should therefore be measured for each individual patient in order to estimate the total number of decays in a particular organ/tissue. The way

  12. Dosimetry and operation of irradiation facilities

    International Nuclear Information System (INIS)

    The industrial use of ionizing radiation has required, from the very first, the measurement of delivered and absorbed doses; hence the necessity of providing dosimetric systems. Laboratories, scientists, industries and potential equipment manufacturers have all collaborated in this new field of activity. Dosimetric intercomparisons have been made by each industry at their own facilities and in collaboration with specialists, national organizations and the IAEA. Dosimetry has become a way of ensuring that treatment by irradiation has been carried out in accordance with the rules. It has become in effect assurance of quality. Routine dosimetry should determine a maximum and minimum dose. Numerous factors play a part in dosimetry. Industry is currently in possession of routine dosimetric systems that are sufficiently accurate, fairly easy to handle and reasonable in cost, thereby satisfying all the requirements of industry and the need for control. Dosimetry is important in the process of marketing irradiated products. The operator of an industrial irradiation facility bases his dosimetry on comparison with reference systems. Research aimed at simplifying the practice of routine dosimetry should be continued. New physical and chemical techniques will be incorporated into systems already in use. The introduction of microcomputers into the operation of radiation facilities has increased the value of dosimetry and made the conditions of treatment more widespread. Stress should be placed on research in several areas apart from reference systems, for example: dosimetric systems at temperatures from +8 deg. C to -45 deg. C, over the dose range 100 krad to a little more than 1 Mrad, liquids and fluidized solids carried at high speed through ducts, thin-film liquids circulating at a high flow rate, and various other problems. (author)

  13. From Takeoff to Landing in Accrediting Corporate Governance Training Programs

    OpenAIRE

    Husseiny, Yehia El

    2009-01-01

    Most of the Organization for Economic Cooperation and Development (OECD) member countries has training institutions that offer corporate governance courses and programs for directors (e.g., the Institute of Directors in the United Kingdom, and the National Association of Corporate Directors in the United States) that have received independent accreditation. There is a high demand among cor...

  14. Internships in School Psychology: Selection and Accreditation Issues

    Science.gov (United States)

    Keilin, W. Gregory

    2015-01-01

    Doctoral students in school psychology often report unique issues and challenges when seeking a doctoral internship. The number and range of accredited internship positions available to School Psychology (SP) students in the Association of Psychology Postdoctoral and Internship Centers (APPIC) Match is quite limited, and they often obtain…

  15. Southern Accrediting Agency Takes a Closer Look at College Sports.

    Science.gov (United States)

    Blum, Debra E.

    1995-01-01

    The Southern Association of Colleges and Schools' recommendation that the University of North Carolina at Chapel Hill place the athletics program solidly in the control of the university, rather than the booster organization, signaled the accrediting agency's willingness to crack down on college sports programs. (MSE)

  16. 9 CFR 77.13 - Accreditation preparatory States or zones.

    Science.gov (United States)

    2010-01-01

    ... TUBERCULOSIS Cattle and Bison § 77.13 Accreditation preparatory States or zones. (a) The following are... livestock other than cattle or bison are included in a newly assembled herd on a premises where a... the “Uniform Methods and Rules—Bovine Tuberculosis Eradication” (January 22, 1999 edition), which...

  17. 9 CFR 77.11 - Modified accredited States or zones.

    Science.gov (United States)

    2010-01-01

    ... TUBERCULOSIS Cattle and Bison § 77.11 Modified accredited States or zones. (a) The following are modified... livestock other than cattle or bison are included in a newly assembled herd on a premises where a... the “Uniform Methods and Rules—Bovine Tuberculosis Eradication” (January 22, 1999, edition), which...

  18. Human Resource Development Planning Based on Accreditation Guidelines.

    Science.gov (United States)

    Dowd, Steven B.

    This paper presents a preliminary human resource development plan for the radiography program sponsored by Lincoln Land Community College (Illinois). The plan is based on the "Essentials and Guidelines of an Accredited Program for the Radiographer," initially adopted in 1944, and most recently revised in 1990, it involves the integration of…

  19. Laboratory quality management system: Road to accreditation and beyond

    Directory of Open Access Journals (Sweden)

    V Wadhwa

    2012-01-01

    Full Text Available This review attempts to clarify the concepts of Laboratory Quality Management System (Lab QMS for a medical testing and diagnostic laboratory in a holistic way and hopes to expand the horizon beyond quality control (QC and quality assurance. It provides an insight on accreditation bodies and highlights a glimpse of existing laboratory practices but essentially it takes the reader through the journey of accreditation and during the course of reading and understanding this document, prepares the laboratory for the same. Some of the areas which have not been highlighted previously include: requirement for accreditation consultants, laboratory infrastructure and scope, applying for accreditation, document preparation. This section is well supported with practical illustrations and necessary tables and exhaustive details like preparation of a standard operating procedure and a quality manual. Concept of training and privileging of staff has been clarified and a few of the QC exercises have been dealt with in a novel way. Finally, a practical advice for facing an actual third party assessment and caution needed to prevent post-assessment pitfalls has been dealt with.

  20. AAFCS Accreditation: From Dream to Reality at Jacksonville State University

    Science.gov (United States)

    Goodwin, Debra K.; Roberts, W. Tim; Boggs, Robbie; Townsel, Kim; Frazier, Jeannie; Marsh, Jill

    2014-01-01

    Accreditation by the American Association of Family and Consumer Sciences (AAFCS) was a long-held dream of the Family and Consumer Sciences (FCS) Unit at Jacksonville State University in Alabama. After more than 6 decades of preparing FCS students for life and the workplace, the FCS Unit resolutely began the journey to the coveted and honored…