WorldWideScience

Sample records for acceptance test procedure

  1. Void fraction instrument acceptance test procedure

    Energy Technology Data Exchange (ETDEWEB)

    Pearce, K.L.

    1994-09-15

    This acceptance test procedure (ATP) was written to test the void fraction instrument (VFI) and verify that the unit is ready for field service. The procedure verifies that the mechanical and electrical features (not specifically addressed in the software ATP) and software alarms are operating as designed.

  2. Void fraction instrument acceptance test procedure

    International Nuclear Information System (INIS)

    This acceptance test procedure (ATP) was written to test the void fraction instrument (VFI) and verify that the unit is ready for field service. The procedure verifies that the mechanical and electrical features (not specifically addressed in the software ATP) and software alarms are operating as designed

  3. Gas characterization system software acceptance test procedure

    International Nuclear Information System (INIS)

    This document details the Software Acceptance Testing of gas characterization systems. The gas characterization systems will be used to monitor the vapor spaces of waste tanks known to contain measurable concentrations of flammable gases

  4. Void fraction instrument acceptance test procedure

    International Nuclear Information System (INIS)

    This document presents the results of the acceptance test for the mechanical and electrical features (not specifically addressed by the software ATP) of the void fraction instrument (VFI). Acceptance testing of the VFI, control console, and decontamination spray assembly was conducted in the 306E building high bay and area adjacent to the facility. The VFI was tested in the horizontal position supported in multiple locations on rolling tables. The control console was located next to the VFI pneumatic control assembly. The VFI system was operated exactly as is expected in the tank farm, with the following exceptions: power was provided from a building outlet and the VFI was horizontal. The testing described in this document verifies that the mechanical and electrical features are operating as designed and that the unit is ready for field service

  5. Project W-049H collection system Acceptance Test Procedure

    Energy Technology Data Exchange (ETDEWEB)

    Carrigan, M.C. [Westinghouse Hanford Co., Richland, WA (United States)

    1994-11-11

    The purpose of this Acceptance Test Procedure (ATP) for the Project W-049H, Treated Effluent Disposal Facility, is to verify that the collection system equipment installed as Pump Station No. 1 (225-W) and Pump Station No. 2 (225-E) have been installed in accordance with the design documents and function as required by the project criteria. This will be a wet test with potable water being introduced into the pump pits to test for leakage. Potable water will also be employed in the testing of the pumps and related mechanical equipment. All Instrument and Control equipment related to the pump stations will be checked electronically with simulated inputs/outputs when actual input/output signals are unavailable. Water from Pump Station 1 will be moved through the TEDF piping system and discharged into the disposal ponds. This will check the proper function of the air/vac valves not tested during construction, and the automated samplers.

  6. Loss on Ignition Furnace Acceptance and Operability Test Procedure

    Energy Technology Data Exchange (ETDEWEB)

    JOHNSON, D.C.

    2000-06-01

    The purpose of this Acceptance Test Procedure and Operability Test Procedure (ATP/OTP)is to verify the operability of newly installed LOI equipment, including a model 1608FL CM{trademark} Furnace, a dessicator, and balance. The operability of the furnace will be verified. The arrangement of the equipment placed in Glovebox 157-3/4 to perform Loss on Ignition (LOI) testing on samples supplied from the Thermal Stabilization line will be verified. In addition to verifying proper operation of the furnace, this ATP/OTP will also verify the air flow through the filters, verify a damper setting to establish and maintain the required differential pressure between the glovebox and the room pressure, and test the integrity of the newly installed HEPA filter. In order to provide objective evidence of proper performance of the furnace, the furnace must heat 15 crucibles, mounted on a crucible rack, to 1000 C, according to a program entered into the furnace controller located outside the glovebox. The glovebox differential pressure will be set to provide the 0.5 to 2.0 inches of water (gauge) negative pressure inside the glovebox with an airflow of 100 to 125 cubic feet per minute (cfm) through the inlet filter. The glovebox inlet Glfilter will he flow tested to ensure the integrity of the filter connections and the efficiency of the filter medium. The newly installed windows and glovebox extension, as well as all disturbed joints, will be sonically tested via ultra probe to verify no leaks are present. The procedure for DOS testing of the filter is found in Appendix A.

  7. Acceptance test procedure for Project W-049H

    Energy Technology Data Exchange (ETDEWEB)

    Buckles, D.I. [Westinghouse Hanford Co., Richland, WA (United States)

    1994-09-29

    The Acceptance Test Procedure (ATP) program for Project W-049H (200 Area Treated Effluent Disposal Facility [TEDF]) covers three activities as follows: (1) Disposal System; (2) Collection System; and (3) Instrumentation and Control System. Each activity has its own ATP. The purpose of the ATPs is to reverify that the systems have been constructed in accordance with the construction documents and to demonstrate that the systems function as required by the Project criteria. The Disposal System ATP covers the testing of the following: disposal line flowmeters, room air temperatures in the Disposal Station Sampling Building, effluent valves and position indicators, disposal pond level monitors, automated sampler, pressure relief valves, and overflow diversion sluice gates. The Collection System ATP covers the testing of the two pump stations and all equipment installed therein. The Instrumentation and Control (I and C) ATP covers the testing of the entire TEDF I and C system. This includes 3 OCS units, modem, and GPLI cabinets in the ETC control room; 2 pump stations; disposal station sampling building; and all LCUs installed in the field.

  8. Test Report for Acceptance Test Procedure for Pumping and Instrumentation Control Skid ''K''

    CERN Document Server

    Johns, B R

    1999-01-01

    This is a Test Report for Acceptance Test Procedure (ATP) HNF-4276. This test report provides the results of the inspection and testing of the new Pumping and Instrumentation Control (PIC) skid designed as ''K''. The ATP was successfully completed. A copy of the completed ATP is in the Appendix of this document.

  9. Test Report for Acceptance Test Procedure for Pumping and Instrumentation Control Skid ''K''

    International Nuclear Information System (INIS)

    This is a Test Report for Acceptance Test Procedure (ATP) HNF-4276. This test report provides the results of the inspection and testing of the new Pumping and Instrumentation Control (PIC) skid designed as ''K''. The ATP was successfully completed. A copy of the completed ATP is in the Appendix of this document

  10. Test Report for Acceptance Test Procedure for Pumping Instrumentation and Control Skid N

    Energy Technology Data Exchange (ETDEWEB)

    KOCH, M.R.

    2000-02-03

    This is a Test Report for Acceptance Test Procedure (ATP) RPP-5489. This test report provides the results of the inspection and testing of the new Pumping Instrumentation and Control (PIC) skid designed as ''N''. The ATP was successfully completed. A copy of the completed ATP is in the Appendix of this document.

  11. Test Report for Acceptance Test Procedure for Pumping Instrumentation and Control Skid L

    Energy Technology Data Exchange (ETDEWEB)

    KOCH, M.R.

    1999-11-09

    This is a Test Report for Acceptance Test Procedure (ATP) RPP-5055. This test report provides the results of the inspection and testing of the new Pumping Instrumentation and Control (PIC) skid designed as ''L''. The ATP was successfully completed. A copy of the completed ATP is in the Appendix of this document.

  12. Acceptance test procedure for the 105-KW isolation barrier leak rate

    International Nuclear Information System (INIS)

    This acceptance test procedure shall be used to: First establish a basin water loss rate prior to installation of the two isolation barriers between the main basin and the discharge chute in K-Basin West. Second, perform an acceptance test to verify an acceptable leakage rate through the barrier seals. This Acceptance Test Procedure (ATP) has been prepared in accordance with CM-6-1 EP 4.2, Standard Engineering Practices

  13. Acceptance test procedure for the 105-KW isolation barrier leak rate

    Energy Technology Data Exchange (ETDEWEB)

    McCracken, K.J.

    1995-05-19

    This acceptance test procedure shall be used to: First establish a basin water loss rate prior to installation of the two isolation barriers between the main basin and the discharge chute in K-Basin West. Second, perform an acceptance test to verify an acceptable leakage rate through the barrier seals. This Acceptance Test Procedure (ATP) has been prepared in accordance with CM-6-1 EP 4.2, Standard Engineering Practices.

  14. Acceptance Test Procedure for New Pumping and Instrumentation Control Skid sup L sup

    CERN Document Server

    Koch, M R

    1999-01-01

    This Acceptance Test Procedure (ATP) provides for the inspection and testing of the new Pumping and Instrumentation Control (PIC) skid designed as ''L''. The ATP will be performed after the construction of the PIC skid in the shop.

  15. PDSS/IMC qualification test software acceptance procedures

    Science.gov (United States)

    1984-01-01

    Tests to be performed for qualifying the payload development support system image motion compensator (IMC) are identified. The performance of these tests will verify the IMC interfaces and thereby verify the qualification test software.

  16. Tank Monitoring and Control System (TMACS) Acceptance Test Procedure

    Energy Technology Data Exchange (ETDEWEB)

    BARNES, D.A.

    2000-06-01

    The purpose of this document is to describe tests performed to validate Revision 12.0 of the TMACS Monitor and Control System (TMACS) and verify that the software functions as intended by design. This document is intended to test the software portion of TMACS. The tests will be performed on the development system. The software to be tested is the TMACS knowledge bases (KB) and the I/O driver/services. The development system will not be communicating to field equipment; instead, the field equipment is simulated using emulators or multiplexers in the lab.

  17. Acceptance Test Procedure for New Pumping Instrumentation and Control Skid ''V''

    International Nuclear Information System (INIS)

    This Acceptance Test Procedure (ATP) verifies proper construction per the design drawings and tests for proper functioning of the Pumping Instrumentation and Control (PIC) skid ''V''. The scope section lists the systems and functions to be checked. This ATP will be performed at the Site Fabrication Services (SFS) shop upon completion of the construction of the PIC skid

  18. Acceptance Test Procedure for New Pumping Instrumentation and Control Skid U

    Energy Technology Data Exchange (ETDEWEB)

    KOCH, M.R.

    2000-12-05

    This Acceptance Test Procedure (ATP) provides for the inspection and testing of the new Pumping Instrumentation and Control (PIC) skid designed as ''U''. The ATP will be performed after the construction of the PIC skid in the fabrication shop.

  19. Test Report for Acceptance Test Procedure for Pumping Instrumentation and Control Skid ''P''

    International Nuclear Information System (INIS)

    This Acceptance Test Report (ATR) provides the test results for the inspection and testing of the new Pumping Instrumentation and Control (PIC) skid designed as ''P''. The ATR summaries the results and provides a copy of the ATP and inspections in the Appendix

  20. Test Report for Acceptance Test Procedure for Pumping Instrumentation and Control Skid P

    Energy Technology Data Exchange (ETDEWEB)

    KOCH, M.R.

    2000-03-29

    This Acceptance Test Report (ATR) provides the test results for the inspection and testing of the new Pumping Instrumentation and Control (PIC) skid designed as ''P''. The ATR summaries the results and provides a copy of the ATP and inspections in the Appendix.

  1. Test Report for Acceptance Test Procedure for Pumping Instrumentation and Control Skid Q

    Energy Technology Data Exchange (ETDEWEB)

    KOCH, M.R.

    2000-05-11

    This Acceptance Test Report (ATR) provides the test results for the inspection and testing of the new Pumping Instrumentation and Control (PIC) skid designed as ''Q''. The ATR summaries the results and provides a copy of the ATP and inspections in the Appendix.

  2. Acceptance test procedure, 241-SY-101/241-C-106 shot loading system

    International Nuclear Information System (INIS)

    This Acceptance Test Procedure is for the 241-SY-101/241-C-106 Shot Loading System. The procedure will test the components of the Shot Loading System and its capability of adequately loading shot into the annular space of the Container. The loaded shot will provide shielding as required for transporting and storage of a contaminated pump after removal from the tank. This test serves as verification that the SLS is acceptable for use in the pump removal operations for Tanks 241-SY-101, 241-C-106 and 241-AY-102. The pump removal operation for these three tanks will be performed by two different organizations with different equipment, but the Shot Loading System will be compatible between the two operations

  3. Acceptance test procedure for K basins dose reduction project clean and coat equipment

    Energy Technology Data Exchange (ETDEWEB)

    Creed, R.F.

    1996-03-11

    This document is the Acceptance Test Procedure (ATP) for the clean and coat equipment designed by Oceaneering Hanford, Inc. under purchase order MDK-XVC-406988 for use in the 105 K East Basin. The ATP provides the guidelines and criteria to test the equipment`s ability to clean and coat the concrete perimeter, divider walls, and dummy elevator pit above the existing water level. This equipment was designed and built in support of the Spent Nuclear Fuel, Dose Reduction Project. The ATP will be performed at the 305 test facility in the 300 Area at Hanford. The test results will be documented in WHC-SD-SNF-ATR-020.

  4. Building Energy Simulation Test for Existing Homes (BESTEST-EX): Instructions for Implementing the Test Procedure, Calibration Test Reference Results, and Example Acceptance-Range Criteria

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, R.; Polly, B.; Bianchi, M.; Neymark, J.; Kennedy, M.

    2011-08-01

    This publication summarizes building energy simulation test for existing homes (BESTEST-EX): instructions for implementing the test procedure, calibration tests reference results, and example acceptance-range criteria.

  5. CENRTC Project {number_sign}2F3EOA, OCB A-376, acceptance test procedure

    Energy Technology Data Exchange (ETDEWEB)

    Akerson, A.W.

    1995-01-25

    This test procedure provides the steps necessary to verify correct functional operation of controls, annunciators, alarms, protective relays and related systems impacted by CENRTC {number_sign}2F3E0A, Microwave Transfer Trip Project, modification work performed under work package 6B-93-00041/M. This procedure separates three tests into separate sections: energization of A-376 duplex panel and circuits; SCADA tests; and A-376 local trip tests.

  6. CENRTC Project {number_sign}2F3EOA, OCB A-386, acceptance test procedure

    Energy Technology Data Exchange (ETDEWEB)

    Akerson, A.W.

    1995-03-06

    This test procedure provides the steps necessary to verify correct functional operation of controls, annunciators, alarms, protective relays and related systems impacted by CENRTC {number_sign}2F3EOA, Microwave Transfer Trip Project, modification work performed under work package 6B-93-00043/M (CENRTC 2F3EOA MWTT OCB A-386 PACKAGE). This procedure separates four tests into separate sections: Energization of A-386 Duplex Panel and Circuits; Local RFL 6750 function tests and start-up; SCADA tests; and A-386 local trip tests.

  7. GENRTC Project No. 2F3E0A, OCB A-382, acceptance test procedure

    Energy Technology Data Exchange (ETDEWEB)

    Akerson, A.W.

    1995-04-10

    This test procedure provides the steps necessary to verify correct functional operation of controls, annunciators, alarms, protective relays and related systems impacted by CENRTC {number_sign}2F3E0A, Microwave Transfer Trip Project, modification work performed under work package 6B-93-00040/M (CENRTC 2F3E0A MWTT OCB A-382 PACKAGE).

  8. CENRTC Project No. 2F3EOA, OCB A-372, acceptance test procedure

    Energy Technology Data Exchange (ETDEWEB)

    Akerson, A.W.

    1995-01-01

    This test procedure provides the steps necessary to verify correct functional operation of controls, annunciators, alarms, protective relays and related systems impacted by CENRTC No. 2F3EOA, Microwave Transfer Trip Project, modification work performed under work package 6B-93-00038/M (CENRTC 2F3EOA MWTT OCB A-372 PACKAGE).

  9. Cone penetrometer acceptance test report

    Energy Technology Data Exchange (ETDEWEB)

    Boechler, G.N.

    1996-09-19

    This Acceptance Test Report (ATR) documents the results of acceptance test procedure WHC-SD-WM-ATR-151. Included in this report is a summary of the tests, the results and issues, the signature and sign- off ATP pages, and a summarized table of the specification vs. ATP section that satisfied the specification.

  10. 105KE Basin Area Radiation Monitor System (ARMS) Acceptance Test Procedure

    CERN Document Server

    Kinkel, C C

    1999-01-01

    This procedure is intended for the Area Radiation Monitoring System, ARMS, that is replacing the existing Programmable Input-Output Processing System, PIOPS, radiation monitoring system in the 105KE basin. The new system will be referred to as the 105KE ARMS, 105KE Area Radiation Monitoring System. This ATP will ensure calibration integrity of the 105KE radiation detector loops. Also, this ATP will test and document the display, printing, alarm output, alarm acknowledgement, upscale check, and security functions. This ATP test is to be performed after completion of the 105KE ARMS installation. The alarm outputs of the 105KE ARMS will be connected to the basin detector alarms, basin annunciator system, and security Alarm Monitoring System, AMS, located in the 200 area Central Alarm Station (CAS).

  11. Human Factors Process Task Analysis Liquid Oxygen Pump Acceptance Test Procedure for the Advanced Technology Development Center

    Science.gov (United States)

    Diorio, Kimberly A.

    2002-01-01

    A process task analysis effort was undertaken by Dynacs Inc. commencing in June 2002 under contract from NASA YA-D6. Funding was provided through NASA's Ames Research Center (ARC), Code M/HQ, and Industrial Engineering and Safety (IES). The John F. Kennedy Space Center (KSC) Engineering Development Contract (EDC) Task Order was 5SMA768. The scope of the effort was to conduct a Human Factors Process Failure Modes and Effects Analysis (HF PFMEA) of a hazardous activity and provide recommendations to eliminate or reduce the effects of errors caused by human factors. The Liquid Oxygen (LOX) Pump Acceptance Test Procedure (ATP) was selected for this analysis. The HF PFMEA table (see appendix A) provides an analysis of six major categories evaluated for this study. These categories include Personnel Certification, Test Procedure Format, Test Procedure Safety Controls, Test Article Data, Instrumentation, and Voice Communication. For each specific requirement listed in appendix A, the following topics were addressed: Requirement, Potential Human Error, Performance-Shaping Factors, Potential Effects of the Error, Barriers and Controls, Risk Priority Numbers, and Recommended Actions. This report summarizes findings and gives recommendations as determined by the data contained in appendix A. It also includes a discussion of technology barriers and challenges to performing task analyses, as well as lessons learned. The HF PFMEA table in appendix A recommends the use of accepted and required safety criteria in order to reduce the risk of human error. The items with the highest risk priority numbers should receive the greatest amount of consideration. Implementation of the recommendations will result in a safer operation for all personnel.

  12. Feasibility of the gamma camera acceptance testing procedure introduced by the Swiss Federal Office of public health

    Energy Technology Data Exchange (ETDEWEB)

    Baechler, S.; Bochud, F.; Verdun, F.R. [University Institute for Radiation Physics, Lausanne (Switzerland); Corminboeuf, F. [Klinik and Poliklinik fur Nuklearmedizin der Universitat, Inselspital, Bern (Switzerland); Linder, R.; Trueb, Ph. [Swiss Federal Office of Public Health, Bern (Switzerland); Malterre, J.; Bischof Delaloye, A. [University Hospital of Lausanne (CHUV), Nuclear Medicine Dept., Lausanne (Switzerland)

    2006-07-01

    Like in the field of radiology, digital systems are also becoming the standard in the field of nuclear medicine. This offers not only the possibility to process, transmit and archive data from patients more easily but also to introduce quantitative measurements for quality controls. In this framework, standards concerning the qualification of gamma camera systems have been updated and appeared to be useful to set legal requirements, in spite of the fact, that this is not their goals. The aim of this study was first to choose a set of tests described in standards to define measurements to be performed at the acceptance of the systems and after the regular maintenance (at least once every six months). Reference values are then established to control the stability of the system. To verify the feasibility, from a technical and a time requirements points of view, the tests proposed for the quality assurance programme have been applied on three gamma camera systems. The results of this study show that new requirements concerning the quality assurance of gamma camera of the Swiss Federal Office of Public Health based on international standards required to slightly modify some procedures to reduce the time necessary for the acceptance and status tests. (authors)

  13. Human Factors Process Task Analysis: Liquid Oxygen Pump Acceptance Test Procedure at the Advanced Technology Development Center

    Science.gov (United States)

    Diorio, Kimberly A.; Voska, Ned (Technical Monitor)

    2002-01-01

    This viewgraph presentation provides information on Human Factors Process Failure Modes and Effects Analysis (HF PFMEA). HF PFMEA includes the following 10 steps: Describe mission; Define System; Identify human-machine; List human actions; Identify potential errors; Identify factors that effect error; Determine likelihood of error; Determine potential effects of errors; Evaluate risk; Generate solutions (manage error). The presentation also describes how this analysis was applied to a liquid oxygen pump acceptance test.

  14. Mini-DACS acceptance test report

    International Nuclear Information System (INIS)

    The attached document is the acceptance test summary report, of the mini-DACS, that was used to test the spare mixer pump for tank 241-SY-101, at the maintenance and storage facility (MASF). It includes a copy of the acceptance test procedure (ATP) and all pertaining ECNS

  15. Cone penetrometer moisture probe acceptance test report

    International Nuclear Information System (INIS)

    This Acceptance Test Report (ATR) documents the results of WHC-SD-WM-ATP-146 (Prototype Cone Penetrometer Moisture Probe Acceptance Test Procedure) and WHC-SD-WM-ATP-145 (Cone Penetrometer Moisture Probe Acceptance Test Procedure). The master copy of WHC-SD-WM-ATP-145 can be found in Appendix A and the master copy of WHC-SD-WM-ATP-146 can be found in Appendix B. Also included with this report is a matrix showing design criteria of the cone penetrometer moisture probe and the verification method used (Appendix C)

  16. TMACS Test Procedure TP011: Panalarm Interface

    International Nuclear Information System (INIS)

    The TMACS Software Test Procedures translate the project's acceptance criteria into test steps. The TMACS Test Plan (WHC-SD-WM-TP-148) is fulfilled when all Test Cases are approved. This Test Procedure tests the TMACS Panalarm Interface functions

  17. Nitrogen trailer acceptance test report

    International Nuclear Information System (INIS)

    This Acceptance Test Report documents compliance with the requirements of specification WHC-S-0249. The equipment was tested according to WHC-SD-WM-ATP-108 Rev.0. The equipment being tested is a portable contained nitrogen supply. The test was conducted at Norco's facility

  18. Nitrogen trailer acceptance test report

    Energy Technology Data Exchange (ETDEWEB)

    Kostelnik, A.J.

    1996-02-12

    This Acceptance Test Report documents compliance with the requirements of specification WHC-S-0249. The equipment was tested according to WHC-SD-WM-ATP-108 Rev.0. The equipment being tested is a portable contained nitrogen supply. The test was conducted at Norco`s facility.

  19. Physics acceptance and QA procedures for IMRT

    International Nuclear Information System (INIS)

    QA program for DMLC. We distinguish between the preliminary preparations for clinical DMLC implementation (i.e., acceptance testing and commissioning), the routine QA of DMLC performance (machine QA), and the verification of patient-specific IM fields using DMLC (treatment QA). Since our initial DMLC implementation in 1995, there have been a number of incremental improvements to our overall QA program. These include methods of mechanical and dosimetric measurements, and software tools developed either by us or by the manufacturer. The use of EPID (electronic portal imaging devise), which potentially provide a more efficient, but certainly not 'cost-free' method, awaits improvements in the speed of data acquisition and in spatial resolution. Our current approach integrates existing methods, combining periodic QA and computer verification, to provide the necessary quality assurance in a safe and efficient manner. Copyright (2001) Australasian College of Physical Scientists and Engineers in Medicine

  20. TMACS Test Procedure TP002: Trending

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure tests the TMACS Trending functions

  1. TMACS Test Procedure TP009: Acromag Driver

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure tests the TMACS Acromag Software Driver (Bridge Code)

  2. TMACS Test Procedure TP007: System administration

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure tests the TMACS System Administration functions

  3. Apollo experience report environmental acceptance testing

    Science.gov (United States)

    Laubach, C. H. M.

    1976-01-01

    Environmental acceptance testing was used extensively to screen selected spacecraft hardware for workmanship defects and manufacturing flaws. The minimum acceptance levels and durations and methods for their establishment are described. Component selection and test monitoring, as well as test implementation requirements, are included. Apollo spacecraft environmental acceptance test results are summarized, and recommendations for future programs are presented.

  4. 14 CFR 151.29 - Procedures: Offer, amendment, and acceptance.

    Science.gov (United States)

    2010-01-01

    ... § 151.29 Procedures: Offer, amendment, and acceptance. (a) Upon approving a project, the Administrator... 14 Aeronautics and Space 3 2010-01-01 2010-01-01 false Procedures: Offer, amendment, and acceptance. 151.29 Section 151.29 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION, DEPARTMENT...

  5. 105-KE Isolation Barrier Leak Rate Acceptance Test Report

    Energy Technology Data Exchange (ETDEWEB)

    McCracken, K.J.

    1995-06-14

    This Acceptance Test Report (ATR) contains the completed and signed Acceptance Procedure (ATP) for the 105-KE Isolations Barrier Leak Rate Test. The Test Engineer`s log, the completed sections of the ATP in the Appendix for Repeat Testing (Appendix K), the approved WHC J-7s (Appendix H), the data logger files (Appendices T and U), and the post test calibration checks (Appendix V) are included.

  6. 105-KE Isolation Barrier Leak Rate Acceptance Test Report

    International Nuclear Information System (INIS)

    This Acceptance Test Report (ATR) contains the completed and signed Acceptance Procedure (ATP) for the 105-KE Isolations Barrier Leak Rate Test. The Test Engineer's log, the completed sections of the ATP in the Appendix for Repeat Testing (Appendix K), the approved WHC J-7s (Appendix H), the data logger files (Appendices T and U), and the post test calibration checks (Appendix V) are included

  7. Acceptance test report for the Westinghouse 100 ton hydraulic trailer

    International Nuclear Information System (INIS)

    The SY-101 Equipment Removal System 100 Ton Hydraulic Trailer was designed and built by KAMP Systems, Inc. Performance of the Acceptance Test Procedure at KAMP's facility in Ontario, California (termed Phase 1 in this report) was interrupted by discrepancies noted with the main hydraulic cylinder. The main cylinder was removed and sent to REMCO for repair while the trailer was sent to Lampson's facility in Pasco, Washington. The Acceptance Test Procedure was modified and performance resumed at Lampson (termed Phase 2 in this report) after receipt of the repaired cylinder. At the successful conclusion of Phase 2 testing the trailer was accepted as meeting all the performance criteria specified

  8. TMACS Test Procedure TP010: Integration summary. Revision 5

    International Nuclear Information System (INIS)

    The TMACS Soft Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes

  9. Nevada Test Site Waste Acceptance Criteria (NTSWAC)

    Energy Technology Data Exchange (ETDEWEB)

    NNSA/NSO Waste Management Project

    2008-06-01

    This document establishes the U.S. Department of Energy, National Nuclear Security Administration Nevada Site Office, Nevada Test Site Waste Acceptance Criteria (NTSWAC). The NTSWAC provides the requirements, terms, and conditions under which the Nevada Test Site will accept low-level radioactive (LLW) and LLW Mixed Waste (MW) for disposal.

  10. Acceptance test report for core sample trucks 3 and 4

    International Nuclear Information System (INIS)

    The purpose of this Acceptance Test Report is to provide documentation for the acceptance testing of the rotary mode core sample trucks 3 and 4, designated as HO-68K-4600 and HO-68K-4647, respectively. This report conforms to the guidelines established in WHC-IP-1026, ''Engineering Practice Guidelines,'' Appendix M, ''Acceptance Test Procedures and Reports.'' Rotary mode core sample trucks 3 and 4 were based upon the design of the second core sample truck (HO-68K-4345) which was constructed to implement rotary mode sampling of the waste tanks at Hanford. Successful completion of acceptance testing on June 30, 1995 verified that all design requirements were met. This report is divided into four sections, beginning with general information. Acceptance testing was performed on trucks 3 and 4 during the months of March through June, 1995. All testing was performed at the ''Rock Slinger'' test site in the 200 West area. The sequence of testing was determined by equipment availability, and the initial revision of the Acceptance Test Procedure (ATP) was used for both trucks. Testing was directed by ICF-KH, with the support of WHC Characterization Equipment Engineering and Characterization Project Operations. Testing was completed per the ATP without discrepancies or deviations, except as noted

  11. Acceptance Test Plan for ANSYS Software

    International Nuclear Information System (INIS)

    This plan governs the acceptance testing of the ANSYS software (Full Mechanical Release 5.5) for use on Project Word Management Contract (PHMC) computer systems (either UNIX or Microsoft Windows/NT). There are two phases to the acceptance testing covered by this test plan: program execution in accordance with the guidance provided in installation manuals; and ensuring results of the execution are consistent with the expected physical behavior of the system being modeled

  12. DACS upgrade acceptance test report

    International Nuclear Information System (INIS)

    The DACS, which is housed in a trailer located just outside of the north fence at the SY tank farm, receives input signals from a variety of sensors located in and around the SY-101 tank. These sensors provide information such as: (1) tank vapor space and ventilation system H2 concentration; (2) tank waste temperature; (3) tank pressure; (4) waste density; (5) operating pump parameters such as speed, flow, rotational position, discharge pressure, and internal temperature; (6) strain (for major equipment); and (7) waste level. The output of these sensors is conditioned and transmitted to the DACS computers where these signals are displayed, recorded, and monitored for out-of-specification conditions. If abnormal conditions are detected, then, in certain situations, the DACS automatically generates alarms and causes the system to abort pump operations. The report documents testing performed per WHC-SD-WM-ATP-082. Rev. 0-13

  13. Acceptance test report, plutonium finishing plant life safety upgrade

    Energy Technology Data Exchange (ETDEWEB)

    Hodge, S.G.

    1994-12-02

    This acceptance Test Procedure (ATP) has been prepared to demonstrate that modifications to the Fir Protection systems function as required by project criteria. The ATP will test the Fire Alarm Control Panels, Flow Alarm Pressure Switch, Heat Detectors, Smoke Detectors, Flow Switches, Manual Pull Stations, and Gong/Door By Pass Switches.

  14. W-026, acceptance test report LLW supercompactor (submittal number 567)

    International Nuclear Information System (INIS)

    The purpose of this acceptance test report was to verify the assembly is correct and complete and meets all the specification and performance requirements. On May 15--16, 1996, the LLW Supercompactor functional demonstrations and performance tests were carried out. These tests were carried out in accordance with the Acceptance Test Procedure, submittal 565, dated 5/3/96. The tests were performed by Herman Miller and Gene Rice of INET Corporation and Richard Lawrence and other PCL personnel. The VIHC representative on May 15, 1996 was Doug Dunlap and on May 16, 1996 the representative was Ken Leist

  15. Gas characterization system software acceptance test report

    International Nuclear Information System (INIS)

    This document details the results of software acceptance testing of gas characterization systems. The gas characterization systems will be used to monitor the vapor spaces of waste tanks known to contain measurable concentrations of flammable gases

  16. Acceptance tests of a new gamma camera

    International Nuclear Information System (INIS)

    For best patient service, a QA programme is needed to produce quantitative/qualitative data and keep records of the results and equipment faults. Gamma cameras must be checked against the manufacturer's specifications.The service manual is usually useful to achieve this goal. Acceptance tests are very important not only to accept a new gamma camera system for routine clinical use but also to have a role in a reference for future measurements. In this study, acceptance tests were performed for a new gamma camera in our department. It is a General Electric MG system with two detectors, two collimators. They are low energy general purpose (LEGP) and medium energy general purpose (MEGP). All intrinsic calibrations and corrections were done by the service engineer at installation (PM tune, dynamic correction, energy calibration, geometric calibration, energy correction, linearity correction and second order corrections).After installation, calibrations and corrections, a close physical inspection of the mechanical and electrical safety aspects of the cameras were done by the responsible physicist of the department. The planar system is based on measurement of system uniformity, resolution/linearity and multiple window spatial registration. All test procedures were performed according to NEMA procedures developed by the manufacturer. Intrinsic uniformity: NEMA uniformity was done first by using service manual and then other isotope uniformities were acquired with 99mTc, 131I, 201Tl and 67Ga. They were evaluated qualitatively and quantitatively, but non-uniformities were observed, especially for detector II, The service engineers repeated all tests and made necessary corrections. We repeated all the intrinsic uniformity tests. 99mTc intrinsic images were also performed at 'no correction', 'no energy correction', 'no linearity correction', 'all correction' and '±10% off peak', and compared. Extrinsic uniformity: At the beginning, collimators were checked for defects

  17. RISK ANALYSES USED IN ACCEPTANCE TESTING

    Directory of Open Access Journals (Sweden)

    Oxana STOROJ

    2016-06-01

    Full Text Available This article is talking about risk based testing approach in user acceptance testing UAT (User Acceptance Testing. There are presented definitions of risk and risk based testing. In addition, we are talking about risks that can appear during UAT and we are describing the process of testing based on risks. We propose some techniques and methods of identifying risks such as using Brainstorming, Delphi method,probability analysis method and others. Also, risk traceability matrix is presented as a method of prioritizing risks.

  18. NEVADA TEST SITE WASTE ACCEPTANCE CRITERIA

    Energy Technology Data Exchange (ETDEWEB)

    U.S. DEPARTMENT OF ENERGY, NATIONAL NUCLEAR SECURITY ADMINISTRATION, NEVADA SITE OFFICE

    2005-07-01

    This document establishes the U. S. Department of Energy, National Nuclear Security Administration Nevada Site Office (NNSA/NSO) waste acceptance criteria (WAC). The WAC provides the requirements, terms, and conditions under which the Nevada Test Site will accept low-level radioactive and mixed waste for disposal. Mixed waste generated within the State of Nevada by NNSA/NSO activities is accepted for disposal. It includes requirements for the generator waste certification program, characterization, traceability, waste form, packaging, and transfer. The criteria apply to radioactive waste received at the Nevada Test Site Area 3 and Area 5 Radioactive Waste Management Site for storage or disposal.

  19. Procedural justice in children: Preschoolers accept unequal resource distributions if the procedure provides equal opportunities.

    Science.gov (United States)

    Grocke, Patricia; Rossano, Federico; Tomasello, Michael

    2015-12-01

    When it is not possible to distribute resources equitably to everyone, people look for an equitable or just procedure. In the current study, we investigated young children's sense of procedural justice. We tested 32 triads of 5-year-olds in a new resource allocation game. Triads were confronted with three unequal reward packages and then agreed on a procedure to allocate them among themselves. To allocate the rewards, they needed to use a "wheel of fortune." Half of the groups played with a fair wheel (where each child had an equal chance of obtaining each reward package), and the other half played with an unfair wheel. We analyzed children's interactions when using the wheel and conducted an interview with each child after the game was over. Children using the unfair wheel often decided to change the rules of the game, and they also rated it as an unfair procedure in the interview. In contrast, children who played with the fair wheel were mostly accepting of both the outcome and the procedure. Overall, we found that children as young as preschool age are already sensitive not only to distributive justice but to procedural justice as well. PMID:26255603

  20. TMACS Test Procedure TP008: SACS Interface. Revision 5

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure tests the TMACS SACS Interface functions

  1. TMACS test procedure TP009: Acromag driver. Revision 5

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addresses the testing of the functionality of the TMACS Acromag driver software in conjunction with a new bridge for the Panalarm Annunciator system

  2. W-026, transuranic waste (TRU) glovebox acceptance test report

    Energy Technology Data Exchange (ETDEWEB)

    Leist, K.J.

    1998-03-11

    On July 18, 1997, the Transuranic (TRU) glovebox was tested using glovebox acceptance test procedure 13021A-86. The primary focus of the glovebox acceptance test was to examine control system interlocks, display menus, alarms, and operator messages. Limited mechanical testing involving the drum ports, hoists, drum lifter, compacted drum lifter, drum tipper, transfer car, conveyors, sorting table, lidder/delidder device and the TRU empty drum compactor were also conducted. As of February 25, 1998, 10 of the 102 test exceptions that affect the TRU glovebox remain open. These items will be tracked and closed via the WRAP Master Test Exception Database. As part of Test Exception resolution/closure the responsible individual closing the Test Exception performs a retest of the affected item(s) to ensure the identified deficiency is corrected, and, or to test items not previously available to support testing. Test exceptions are provided as appendices to this report.

  3. WRAP low level waste (LLW) glovebox acceptance test report

    Energy Technology Data Exchange (ETDEWEB)

    Leist, K.J.

    1998-02-17

    In June 28, 1997, the Low Level Waste (LLW) glovebox was tested using glovebox acceptance test procedure 13031A-85. The primary focus of the glovebox acceptance test was to examine control system interlocks, display menus, alarms, and operator messages. Limited mechanical testing involving the drum ports, hoists, drum lifter, compacted drum lifter, drum tipper, transfer car, conveyors, lidder/delidder device and the supercompactor were also conducted. As of November 24, 1997, 2 of the 131 test exceptions that affect the LLW glovebox remain open. These items will be tracked and closed via the WRAP Master Test Exception Database. As part of Test Exception resolution/closure the responsible individual closing the Test Exception performs a retest of the affected item(s) to ensure the identified deficiency is corrected, and, or to test items not previously available to support testing. Test Exceptions are provided as appendices to this report.

  4. W-026, transuranic waste (TRU) glovebox acceptance test report

    International Nuclear Information System (INIS)

    On July 18, 1997, the Transuranic (TRU) glovebox was tested using glovebox acceptance test procedure 13021A-86. The primary focus of the glovebox acceptance test was to examine control system interlocks, display menus, alarms, and operator messages. Limited mechanical testing involving the drum ports, hoists, drum lifter, compacted drum lifter, drum tipper, transfer car, conveyors, sorting table, lidder/delidder device and the TRU empty drum compactor were also conducted. As of February 25, 1998, 10 of the 102 test exceptions that affect the TRU glovebox remain open. These items will be tracked and closed via the WRAP Master Test Exception Database. As part of Test Exception resolution/closure the responsible individual closing the Test Exception performs a retest of the affected item(s) to ensure the identified deficiency is corrected, and, or to test items not previously available to support testing. Test exceptions are provided as appendices to this report

  5. WRAP low level waste (LLW) glovebox acceptance test report

    International Nuclear Information System (INIS)

    In June 28, 1997, the Low Level Waste (LLW) glovebox was tested using glovebox acceptance test procedure 13031A-85. The primary focus of the glovebox acceptance test was to examine control system interlocks, display menus, alarms, and operator messages. Limited mechanical testing involving the drum ports, hoists, drum lifter, compacted drum lifter, drum tipper, transfer car, conveyors, lidder/delidder device and the supercompactor were also conducted. As of November 24, 1997, 2 of the 131 test exceptions that affect the LLW glovebox remain open. These items will be tracked and closed via the WRAP Master Test Exception Database. As part of Test Exception resolution/closure the responsible individual closing the Test Exception performs a retest of the affected item(s) to ensure the identified deficiency is corrected, and, or to test items not previously available to support testing. Test Exceptions are provided as appendices to this report

  6. Acceptance test report: Backup power system

    International Nuclear Information System (INIS)

    Acceptance Test Report for construction functional testing of Project W-030 Backup Power System. Project W-030 provides a ventilation upgrade for the four Aging Waste Facility tanks. Backup power includes a single 125 KW diesel generator, three 10-kva uninterruptible power supply units, and all necessary control

  7. User Acceptance of YouTube for Procedural Learning: An Extension of the Technology Acceptance Model

    Science.gov (United States)

    Lee, Doo Young; Lehto, Mark R.

    2013-01-01

    The present study was framed using the Technology Acceptance Model (TAM) to identify determinants affecting behavioral intention to use YouTube. Most importantly, this research emphasizes the motives for using YouTube, which is notable given its extrinsic task goal of being used for procedural learning tasks. Our conceptual framework included two…

  8. AZ-101 Mixer Pump Test Qualification Test Procedures (QTP)

    International Nuclear Information System (INIS)

    Describes the Qualification test procedure for the AZ-101 Mixer Pump Data Acquisition System (DAS). The purpose of this Qualification Test Procedure (QTP) is to confirm that the AZ-101 Mixer Pump System has been properly programmed and hardware configured correctly. This QTP will test the software setpoints for the alarms and also check the wiring configuration from the SIMcart to the HMI. An Acceptance Test Procedure (ATP), similar to this QTP will be performed to test field devices and connections from the field

  9. Tests and Procedures

    Science.gov (United States)

    ... procedure is being done. How the results will influence treatment. What your child will experience during the ... Understanding Children’s Cancer About Cancer What is Cancer? Childhood Cancer Statistics Childhood Cancer Statistics Overview Number of ...

  10. GERDA phase II detectors and acceptance tests

    International Nuclear Information System (INIS)

    The GERDA collaboration aims at searching for the neutrinoless double beta decay of 76Ge into 76Se. It uses enriched Ge detectors that are operated in liquid argon. Phase I of the experiment started with a target mass of approximately 20 kg on November 1, 2011, and it will last for 1 year. Phase II is in preparation and envisions the installation of at least 25 new Broad Energy Ge detectors (BEGe) that will increase the target mass by nearly 20 kg. These detectors, however, have to be extensively tested prior to their usage. The talk briefly reviews the preparation of the detector screening campaign and focuses on the corresponding acceptance test strategy. The tests include energy resolution and leakage current measurements, dead layer and active volume determination, as well as the pulse shape discrimination efficiencies of the detectors. Finally we present results obtained with depleted BEGe detectors that are used for the optimization of the enriched ones during the acceptance tests.

  11. Acceptance Testing Of Web Applications With Test Description Language

    OpenAIRE

    Łukasz Olek; Bartosz Alchimowicz; Jerzy Nawrocki

    2014-01-01

    Acceptance tests are usually created by a client after a part of a system is implemented. However, some methodologies propose the elaboration of test cases before implementing a system. This approach increases the probability of system implementation that fulfills requirements, but may be problematic for customers and testers. To allow acceptance testing in such conditions, we propose to define test cases by recording them on an interactive mockup (a low detailed user-interface prototype). Th...

  12. Nevada Test Site Waste Acceptance Criteria

    Energy Technology Data Exchange (ETDEWEB)

    U. S. Department of Energy, National Nuclear Security Administration Nevada Site Office

    2005-10-01

    This document establishes the U.S. Department of Energy (DOE), National Nuclear Security Administration Nevada Site Office (NNSA/NSO) waste acceptance criteria (WAC). The WAC provides the requirements, terms, and conditions under which the Nevada Test Site (NTS) will accept low-level radioactive (LLW) and mixed waste (MW) for disposal. It includes requirements for the generator waste certification program, characterization, traceability, waste form, packaging, and transfer. The criteria apply to radioactive waste received at the NTS Area 3 and Area 5 Radioactive Waste Management Complex (RWMC) for storage or disposal.

  13. Specification and acceptance testing of radiotherapy treatment planning systems

    International Nuclear Information System (INIS)

    Quality assurance (QA) in the radiation therapy treatment planning process is essential to ensure accurate dose delivery to the patient and to minimize the possibility of accidental exposure. The computerized radiotherapy treatment planning systems (RTPSs) are now widely available in industrialized and developing countries and it is of special importance to support hospitals in Member States in developing procedures for acceptance testing, commissioning and QA of their RTPSs. Responding to these needs, a group of experts developed an IAEA publication with such recommendations, which was published in 2004 as IAEA Technical Reports Series No. 430. This report provides a general framework and describes a large number of tests and procedures that should be considered by the users of new RTPSs. However, small hospitals with limited resources or large hospitals with high patient load and limited staff are not always able to perform complete characterization, validation and software testing of algorithms used in RTPSs. Therefore, the IAEA proposed more specific guidelines that provide a step-by-step recommendation for users at hospitals or cancer centres how to implement acceptance and commissioning procedures for newly purchased RTPSs. The current publication was developed in the framework of the Coordinated Research Project on Development of Procedures for Quality Assurance for Dosimetry Calculations in Radiotherapy and uses the International Electrotechnical Commission (IEC) standard IEC 62083, Requirements for the Safety of Radiotherapy Treatment Planning Systems as its basis. The report addresses the procedures for specification and acceptance testing of RTPSs to be used by both manufacturers and users at the hospitals. Recommendations are provided for specific tests to be performed at the manufacturing facility known as type tests, and for acceptance tests to be performed at the hospital known as site tests. The purpose of acceptance testing is to demonstrate to the

  14. TMACS test procedure TP010: Integration summary. Revision 6

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure verifies that Test Procedures 1, 2, 3, 5 and 9 (WHC-SD-WM-TRP-105, 106, 107, 109 and 113) of TMACS Software Release 4.1 have been successfully completed

  15. Acceptance Testing Of Web Applications With Test Description Language

    Directory of Open Access Journals (Sweden)

    Łukasz Olek

    2014-01-01

    Full Text Available Acceptance tests are usually created by a client after a part of a system is implemented. However, some methodologies propose the elaboration of test cases before implementing a system. This approach increases the probability of system implementation that fulfills requirements, but may be problematic for customers and testers. To allow acceptance testing in such conditions, we propose to define test cases by recording them on an interactive mockup (a low detailed user-interface prototype. The paper focuses on Test Description Language, a notation used to store test cases.

  16. Acceptance test report: Field test of mixer pump for 241-AN-107 caustic addition project

    International Nuclear Information System (INIS)

    The field acceptance test of a 75 HP mixer pump (Hazleton serial number N-20801) installed in Tank 241-AN-107 was conducted from October 1995 thru February 1996. The objectives defined in the acceptance test were successfully met, with two exceptions recorded. The acceptance test encompassed field verification of mixer pump turntable rotation set-up and operation, verification that the pump instrumentation functions within established limits, facilitation of baseline data collection from the mixer pump mounted ultrasonic instrumentation, verification of mixer pump water flush system operation and validation of a procedure for its operation, and several brief test runs (bump) of the mixer pump

  17. Site acceptance test, W-030 MICON system

    Energy Technology Data Exchange (ETDEWEB)

    Hill, L.F., Westinghouse Hanford

    1996-06-10

    Monitoring and control of the W-030 ventilation upgrade is provided by a distributed control system (DCS) furnished by MICON Corporation. After shipment to the Hanford Site, the site acceptance test (SAT) for this system was conducted in a laboratory environment over a six month period, involving four distinct phases and numerous hardware and software modifications required to correct test exceptions. The final results is a system which is not fully compliant with procurement specifications but is determined to meet minimum Project W-030 safety and functional requirements. A negotiated settlement was reached with the supplier to establish a `path forward` for system implementation. This report documents the `as-run` status of the SAT. The SAT was completed in August of 1995. It was later followed by comprehensive acceptance testing of the W-030 control-logic configuration software; results are documented in WHC-SD-W030-ATR-011. Further testing is reported as part of process system startup operational testing, performed after the MICON installation.

  18. Project W-049H Collection System Acceptance Test

    Energy Technology Data Exchange (ETDEWEB)

    Buckles, D.I.

    1994-11-11

    The Acceptance Test Procedure (ATP) Program for Project W-049H covers the following activities: Disposal system, Collection system, Instrumentation and control system. Each activity has its own ATP. The purpose of the ATPs is to verify that the systems have been constructed in accordance with the construction documents and to demonstrate that the systems function as required by the Project criteria. This ATP has been prepared to demonstrate that the Collection System Instrumentation functions as required by project criteria.

  19. Acceptance Test Plan for Fourth-Generation Corrosion Monitoring Cabinet

    International Nuclear Information System (INIS)

    This Acceptance Test Plan (ATP) will document the satisfactory operation of the third-generation corrosion monitoring cabinet (Hiline Engineering Part No.0004-CHM-072-C01). This ATP will be performed by the manufacturer of the cabinet prior to delivery to the site. The objective of this procedure is to demonstrate and document the acceptance of the corrosion monitoring cabinet. The test will consist of a continuity test of the cabinet wiring from the end of cable to be connected to corrosion probe, through the appropriate intrinsic safety barriers and out to the 15 pin D-shell connectors to be connected to the corrosion monitoring instrument. Additional testing will be performed using a constant current and voltage source provided by the corrosion monitoring hardware manufacturer to verify proper operation of corrosion monitoring instrumentation

  20. 242A Distributed Control System Year 2000 Acceptance Test Report

    International Nuclear Information System (INIS)

    This report documents acceptance test results for the 242-A Evaporator distributive control system upgrade to D/3 version 9.0-2 for year 2000 compliance. This report documents the test results obtained by acceptance testing as directed by procedure HNF-2695. This verification procedure will document the initial testing and evaluation of the potential 242-A Distributed Control System (DCS) operating difficulties across the year 2000 boundary and the calendar adjustments needed for the leap year. Baseline system performance data will be recorded using current, as-is operating system software. Data will also be collected for operating system software that has been modified to correct year 2000 problems. This verification procedure is intended to be generic such that it may be performed on any D/3(trademark) (GSE Process Solutions, Inc.) distributed control system that runs with the VMSTM (Digital Equipment Corporation) operating system. This test may be run on simulation or production systems depending upon facility status. On production systems, DCS outages will occur nine times throughout performance of the test. These outages are expected to last about 10 minutes each

  1. 242A Distributed Control System Year 2000 Acceptance Test Report

    Energy Technology Data Exchange (ETDEWEB)

    TEATS, M.C.

    1999-08-31

    This report documents acceptance test results for the 242-A Evaporator distributive control system upgrade to D/3 version 9.0-2 for year 2000 compliance. This report documents the test results obtained by acceptance testing as directed by procedure HNF-2695. This verification procedure will document the initial testing and evaluation of the potential 242-A Distributed Control System (DCS) operating difficulties across the year 2000 boundary and the calendar adjustments needed for the leap year. Baseline system performance data will be recorded using current, as-is operating system software. Data will also be collected for operating system software that has been modified to correct year 2000 problems. This verification procedure is intended to be generic such that it may be performed on any D/3{trademark} (GSE Process Solutions, Inc.) distributed control system that runs with the VMSTM (Digital Equipment Corporation) operating system. This test may be run on simulation or production systems depending upon facility status. On production systems, DCS outages will occur nine times throughout performance of the test. These outages are expected to last about 10 minutes each.

  2. Acceptance Test Report for the 241-AZ-101 Ultrasonic Interface Level Analyzer

    International Nuclear Information System (INIS)

    This document comprises the Acceptance Test Report for the 241-AZ-101 Ultrasonic Interface Level Analyzer. This document presents the results of Acceptance Testing of the 241-AZ-101 Ultrasonic Interface Level Analyzers (URSILLAs). Testing of the URSILLAs was performed in accordance with ATP-260-001, ''URSILLA Pre-installation Acceptance Test Procedure''. The objective of the testing was to verify that all equipment and components function in accordance with design specifications and original equipment manufacturer's specifications

  3. TMACS test procedure TP004: Reporting. Revision 4

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addressed the report generation capability of the TMACS. The features to be tested are generation of the alarm history report, equipment failure report, single-shell tank report and double-shell tank report

  4. TMACS test procedure TP003: Graphics. Revision 6

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addresses the graphics requirements of the TMACS. The features to be tested are the data display graphics and the graphic elements that provide for operator control and selection of displays

  5. TMACS test procedure TP001: Alarm management. Revision 6

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addresses the Alarm Management requirements of the TMACS. The features to be tested are: real-time alarming on high and low level and discrete alarms, equipment alarms, dead-band filtering, alarm display color coding, alarm acknowledgement, and alarm logging

  6. TMACS test procedure TP003: Graphics. Revision 5

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addresses the graphics requirements of the TMACS. The features to be tested are the data display graphics and the graphic elements that provide for operator control and selection of displays

  7. System turnover and acceptance procedure of Qingshan NPP

    International Nuclear Information System (INIS)

    Turnover and acceptance process of nuclear power plant has direct influence on construction scheduling. The authors describe organization, conditions and implementation of system turnover and acceptance. The paper also gives some suggestions as reference of turnover and acceptance work

  8. Inspection system performance test procedure

    International Nuclear Information System (INIS)

    This procedure establishes requirements to administer a performance demonstration test. The test is to demonstrate that the double-shell tank inspection system (DSTIS) supplied by the contractor performs in accordance with the WHC-S-4108, Double-Shell Tank Ultrasonic Inspection Performance Specification, Rev. 2-A, January, 1995. The inspection system is intended to provide ultrasonic (UT) and visual data to determine integrity of the Westinghouse Hanford Company (WHC) site underground waste tanks. The robotic inspection system consists of the following major sub-systems (modules) and components: Mobile control center; Deployment module; Cable management assembly; Robot mechanism; Ultrasonic testing system; Visual testing system; Pneumatic system; Electrical system; and Control system

  9. Acceptance testing and QA in dental radiology

    International Nuclear Information System (INIS)

    The HPA (UK) reported in 2002 that although the effective dose from dental examinations is low, dental imaging is the most frequent type of X ray examination undertaken (approx. 30%); a similar situation is likely to exist in Ireland. It is most likely that this will increase further in the future as the dental X ray technology moves from film to digital equipment. Dental X rays are taken more frequently with digital equipment than film based technology. The Department of Medical Physics and Bioengineering, St. James's Hospital, Dublin is particularly interested in the area of dental quality assurance (QA) as a result of a significant medical radiation safety incident in Ireland occurring with a faulty dental X ray installation. The implementation of a QA programme in Ireland has been enforced not only by the publication of the Dental Code of Practice by the Irish Regulatory Authority in the 1990s but also by the publication of the European Medical Exposures Directive. The current QA programme, which includes site visits in order to perform acceptance testing and routine QA checks on dental X ray equipment has been implemented in approximately 250 dental health centres to which the Department has provided radiation protection advisory services for approximately twenty years. In addition to the technical evaluation of the equipment, an assessment of the adequacy of the shielding in the X ray room and of the radiation protection facilities is performed. This paper identifies the findings from the dental radiology QA assessments. QA test protocols were developed in accordance with current national and internationally accepted guidelines. The tests include exposure time and kVp accuracy, radiation output consistency, filtration, irradiated field size, mechanical and electrical safety checks and measurement of tube output for specific X ray examinations to aid in the establishment of a national diagnostic reference levels, a requirement of the Medical Exposures

  10. 14 CFR 151.123 - Procedures: Offer; amendment; acceptance; advance planning agreement.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 3 2010-01-01 2010-01-01 false Procedures: Offer; amendment; acceptance... Planning and Engineering Proposals § 151.123 Procedures: Offer; amendment; acceptance; advance planning.... FAA's offer and the sponsor's acceptance constitute an advance planning grant agreement between...

  11. Acceptance and Divergence from Engineering Design Procedures Implicating Knowledge Flow

    DEFF Research Database (Denmark)

    Jensen, Ole Kjeldal; Ahmed, Saeema

    2009-01-01

    of explicit procedures and; 3) implicit procedures supporting needs that are not catered for by the explicit procedures. In this understanding, a procedure can be any kind of method, tool or framework used to support design engineers. Furthermore, the study discusses a variety of recommended actions...

  12. TMACS Test Procedure TP005: Sensor configuration, logging, and data conversion

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the projects acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes

  13. Acceptance test plan for fourth generation Hanford corrosion monitoring system

    International Nuclear Information System (INIS)

    This Acceptance Test Plan (ATP) will document the satisfactory operation of the corrosion probe cabinets destined for installation on tanks 241-AN-102 and 241-AN-107. This ATP will be performed by the manufacturer on each cabinet prior to delivery to the site. The objective of this procedure is to demonstrate and document the acceptance of the corrosion monitoring cabinets to be installed on tanks 241-AN-102 and 241-AN-107. One cabinet will be installed on each tank. Each cabinet will contain corrosion monitoring hardware to be connected to existing corrosion probes already installed in each tank. The test will consist of a continuity test of the cabinet wiring from the end of cable to be connected to corrosion probe, through the appropriate intrinsic safety barriers and out to the 15 pin D-shell connectors to be connected to the corrosion monitoring instrument. Additional testing will be performed using a constant current and voltage source provided by the corrosion monitoring hardware manufacturer to verify proper operation of corrosion monitoring instrumentation (input a known signal and see if the instrumentation records the proper value)

  14. Acceptance Test Report for Fourth Generation Hanford Corrosion Monitoring System

    International Nuclear Information System (INIS)

    This Acceptance Test Report (ATR) will document the satisfactory operation of the corrosion probe cabinets destined for installation on tanks 241-AN-102 and 241-AN-107. This ATR will be performed by the manufacturer on each cabinet prior to delivery to the site. The objective of this procedure is to demonstrate and document the acceptance of the corrosion monitoring cabinets to be installed on tanks 241-AN-102 and 241-AN-107. One cabinet will be installed on each tank. Each cabinet will contain corrosion monitoring hardware to be connected to existing corrosion probes already installed in each tank. The test will consist of a continuity test of the cabinet wiring from the end of cable to be connected to corrosion probe, through the appropriate intrinsic safety barriers and out to the 15 pin D-shell connectors to be connected to the corrosion monitoring instrument. Additional testing will be performed using a constant current and voltage source provided by the corrosion monitoring hardware manufacturer to verify proper operation of corrosion monitoring instrumentation (input a known signal and see if the instrumentation records the proper value)

  15. Probabilistic Requirements (Partial) Verification Methods Best Practices Improvement. Variables Acceptance Sampling Calculators: Empirical Testing. Volume 2

    Science.gov (United States)

    Johnson, Kenneth L.; White, K. Preston, Jr.

    2012-01-01

    The NASA Engineering and Safety Center was requested to improve on the Best Practices document produced for the NESC assessment, Verification of Probabilistic Requirements for the Constellation Program, by giving a recommended procedure for using acceptance sampling by variables techniques as an alternative to the potentially resource-intensive acceptance sampling by attributes method given in the document. In this paper, the results of empirical tests intended to assess the accuracy of acceptance sampling plan calculators implemented for six variable distributions are presented.

  16. User acceptance testing a step-by-step guide

    CERN Document Server

    Hambling, Brian

    2013-01-01

    Every information system brought into service in every type of organisation requires user acceptance testing. This book is a hands-on manual for non-testing specialists to plan and carry out an effective acceptance test of an information system. It also identifies ways of making the process as simple and cost-effective as possible.

  17. Nevada Test Site Waste Acceptance Criteria, December 2000

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2000-12-01

    This document establishes the US Department of Energy, Nevada Operations Office waste acceptance criteria. The waste acceptance criteria provides the requirements, terms, and conditions under which the Nevada Test Site will accept low-level radioactive waste and mixed waste for disposal. It includes requirements for the generator waste certification program, characterization, traceability, waste form, packaging, and transfer. The criteria apply to radioactive waste received at the Nevada Test Site Area 3 and Area 5 Radioactive Waste Management Sites for storage or disposal.

  18. Nevada Test Site Waste Acceptance Criteria, December 2000

    International Nuclear Information System (INIS)

    This document establishes the US Department of Energy, Nevada Operations Office waste acceptance criteria. The waste acceptance criteria provides the requirements, terms, and conditions under which the Nevada Test Site will accept low-level radioactive waste and mixed waste for disposal. It includes requirements for the generator waste certification program, characterization, traceability, waste form, packaging, and transfer. The criteria apply to radioactive waste received at the Nevada Test Site Area 3 and Area 5 Radioactive Waste Management Sites for storage or disposal

  19. Test Plan and Test Procedures Document - Prague Phase II

    OpenAIRE

    Jakobi, Jörn

    2003-01-01

    This document, D16a-TPP, is the output of BETA WP5100 and describes the specific test procedures for Prague airport. It is one of three parts of the “Test Plan and Test Procedures” series of documents. A document is available for each of the test airports to be used in the BETA project: • D16a-TPP Test Plan and Test Procedures document, test procedures for Prague (PRG). • D16b-TPP Test Plan and Test Procedures document, test procedures for Hamburg (HAM). • D16c-TPP Test Plan and Tes...

  20. WRAP low level waste restricted waste management (LLW RWM) glovebox acceptance test report

    Energy Technology Data Exchange (ETDEWEB)

    Leist, K.J.

    1997-11-24

    On April 22, 1997, the Low Level Waste Restricted Waste Management (LLW RWM) glovebox was tested using acceptance test procedure 13027A-87. Mr. Robert L. Warmenhoven served as test director, Mr. Kendrick Leist acted as test operator and test witness, and Michael Lane provided miscellaneous software support. The primary focus of the glovebox acceptance test was to examine glovebox control system interlocks, operator Interface Unit (OIU) menus, alarms, and messages. Basic drum port and lift table control sequences were demonstrated. OIU menus, messages, and alarm sequences were examined, with few exceptions noted. Barcode testing was bypassed, due to the lack of installed equipment as well as the switch from basic reliance on fixed bar code readers to the enhanced use of portable bar code readers. Bar code testing was completed during performance of the LLW RWM OTP. Mechanical and control deficiencies were documented as Test Exceptions during performance of this Acceptance Test. These items are attached as Appendix A to this report.

  1. WRAP low level waste restricted waste management (LLW RWM) glovebox acceptance test report

    International Nuclear Information System (INIS)

    On April 22, 1997, the Low Level Waste Restricted Waste Management (LLW RWM) glovebox was tested using acceptance test procedure 13027A-87. Mr. Robert L. Warmenhoven served as test director, Mr. Kendrick Leist acted as test operator and test witness, and Michael Lane provided miscellaneous software support. The primary focus of the glovebox acceptance test was to examine glovebox control system interlocks, operator Interface Unit (OIU) menus, alarms, and messages. Basic drum port and lift table control sequences were demonstrated. OIU menus, messages, and alarm sequences were examined, with few exceptions noted. Barcode testing was bypassed, due to the lack of installed equipment as well as the switch from basic reliance on fixed bar code readers to the enhanced use of portable bar code readers. Bar code testing was completed during performance of the LLW RWM OTP. Mechanical and control deficiencies were documented as Test Exceptions during performance of this Acceptance Test. These items are attached as Appendix A to this report

  2. Acceptance testing and commissioning of a new model HDR afterloader

    International Nuclear Information System (INIS)

    We have recently performed acceptance testing procedures and have commissioned a new model HDR afterloader, the Varian VariSource with ''Intelligent Drive.'' Our site was one of the first installations worldwide. It is our intent to describe our tests and the results of the tests particularly as they may differ from other afterloaders. The Ir-192 source is unique among afterloaders marketed in the US in that it is very slender (OD of source wire is 0.59 mm) and relatively long (two 0.5 cm sources for a total active length of 1.0 cm). A check of source homogeneity by autoradiograph as urged by the US Nuclear Regulatory Commission demonstrates no detectable source inhomogeneity. Reentrant well ionization chambers are calibrated in the US with a 3.5 mm long source at Accredited Dosimetry Calibration Laboratories. Therefore calibration needs to be considered with some care. Calibration of the first delivered source with a well ionization chamber indicated agreement with the manufacturer's stated activity to within 0.5%. Source positioning is checked with a device called a 'cam scale'. Tests have been carried out on this system and it has been found to accurately indicate source position to within ±0.5 mm. Timer accuracy has been found to be better than 0.1% for dwell times of several hundred seconds. The intelligent drive system and the small source diameter allow the source wire to negotiate paths with small radius of curvature. A series of tests have been made in which the source is forced to negotiate 'U' turns of decreasing radius of curvature. A 4.7 F, 100 cm long catheter was used for these tests and the 'U' turn was positioned at approximately 90 cm. Under these conditions, the VariSource was consistently able to traverse a 1.25 cm radius of curvature, which is better than the manufacturer's stated limit of 1.5 cm

  3. 40 CFR 205.54 - Test procedures.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 24 2010-07-01 2010-07-01 false Test procedures. 205.54 Section 205.54 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) NOISE ABATEMENT PROGRAMS TRANSPORTATION EQUIPMENT NOISE EMISSION CONTROLS Medium and Heavy Trucks § 205.54 Test procedures. The procedures...

  4. Example Procedures for Developing Acceptance-Range Criteria for BESTEST-EX

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, Ron [National Renewable Energy Lab. (NREL), Golden, CO (United States); Polly, Ben [National Renewable Energy Lab. (NREL), Golden, CO (United States); Bianchi, Marcus [National Renewable Energy Lab. (NREL), Golden, CO (United States); Neymark, Joel [J. Neymark & Associates, Golden, CO (United States)

    2010-08-01

    This document provides an example procedure for establishing acceptance-range criteria to assess results from software undergoing BESTEST-EX. This example method for BESTEST-EX is a modified version of the method described in HERS BESTEST.

  5. 16 CFR 1616.4 - Sampling and acceptance procedures.

    Science.gov (United States)

    2010-01-01

    ... garment production unit, provided such colors or print patterns demonstrate char lengths that are not... included in a single garment production unit provided the other garment characteristics are identical... requirement of § 1616.5(c)(4) is met on subsequent fabric production units if results of testing an...

  6. Summary of functional and performance test procedures

    OpenAIRE

    Mitzel, Jens; Gülzow, Erich; Friedrich, K. Andreas; Araya, Samuel; Kaer, Soeren

    2015-01-01

    The EU-funded research project Stack-Test has developed different test procedures for the functional and performance characterization of PEMFC stacks. The different test procedures are pointed out in the current form and the example Performance Mapping Test program is introduced.

  7. Adolescent test or procedure preparation

    Science.gov (United States)

    ... else) during the procedure Playing hand-held video games Using guided imagery Trying other distractions, such as ... Alexander M. Managing patient stress in pediatric radiology. Radiol Technol . ... . Bray L, Callery P, Kirk S. A qualitative study ...

  8. Adolescent test or procedure preparation

    Science.gov (United States)

    ... someone else) during the procedure Playing hand-held video games Using guided imagery Trying other distractions, such as listening to music through headphones, if allowed When possible, let your adolescent make some decisions, such as deciding the time ...

  9. A general design for energy test procedures

    International Nuclear Information System (INIS)

    Appliances are increasingly controlled by microprocessors. Unfortunately, energy test procedures have not been modified to capture the positive and negative contributions of the microprocessor to the appliance's energy use. A new test procedure is described which captures both the mechanical and logical features present in many new appliances. We developed an energy test procedure for refrigerators that incorporates most aspects of the proposed new approach. Some of the strengths and weaknesses of the new test are described

  10. Product acceptance environmental and destructive testing for reliability.

    Energy Technology Data Exchange (ETDEWEB)

    Dvorack, Michael A.; Kerschen, Thomas J.; Collins, Elmer W.

    2007-08-01

    To determine whether a component is meeting its reliability requirement during production, acceptance sampling is employed in which selected units coming off the production line are subjected to additional environmental and/or destructive tests that are within the normal environment space to which the component is expected to be exposed throughout its life in the Stockpile. This report describes what these tests are and how they are scored for reliability purposes. The roles of screens, Engineering Use Only tests, and next assembly product acceptance testing are also discussed, along with both the advantages and disadvantages of environmental and destructive testing.

  11. Handbook of mutagenicity test procedures

    International Nuclear Information System (INIS)

    This book consists of 39 chapters. Some of the titles are: Bacillus subtilis repair test, Induced reversion using human adenovirus, The fluctuation test in bacteria, Chemical mutagenesis with diploid human fibroblasts, The specific locus test in the mouse, The bone marrow micronucleus test, and Sperm morphology in testing in mice

  12. Two Types of Test Assembly Checked and Accepted for CEFR

    Institute of Scientific and Technical Information of China (English)

    2002-01-01

    The detail design of two types of test assembly used in the simulating transportation, and themounting and hydraulic test were completed in 2002. This two types of test assembly are manufactured inshanghai No.1 machine tool works, and checked and accepted in March 2003. The former is used in the

  13. Test report - caustic addition system operability test procedure

    International Nuclear Information System (INIS)

    This Operability Test Report documents the test results of test procedure WHC-SD-WM-OTP-167 ''Caustic Addition System Operability Test Procedure''. The Objective of the test was to verify the operability of the 241-AN-107 Caustic Addition System. The objective of the test was met

  14. Toddler test or procedure preparation

    Science.gov (United States)

    ... the procedure for your child and identify any anxiety your child may have. Tailor this technique to your child. ... with your child for comfort. Avoid showing your anxiety. This will only make your child more nervous. Research suggests that children are more ...

  15. Preschooler test or procedure preparation

    Science.gov (United States)

    ... the procedure for your child and identify any anxiety your child may have. Tailor this technique to your child. ... with your child for comfort. Avoid showing your anxiety. This will only make your child feel more upset. Research suggests that children are ...

  16. 16 CFR 1610.6 - Test procedure.

    Science.gov (United States)

    2010-01-01

    ... FLAMMABILITY OF CLOTHING TEXTILES The Standard § 1610.6 Test procedure. The test procedure is divided into two... be tested is a plain surface textile fabric or a raised surface textile fabric as defined in § 1610.2... flocked overall are treated as raised surface textile fabrics as defined in § 1610.2(l). Flock...

  17. Type of Speech Material Affects Acceptable Noise Level Test Outcome

    OpenAIRE

    Koch, Xaver; Dingemanse, Gertjan; Goedegebure, André; Janse, Esther

    2016-01-01

    The acceptable noise level (ANL) test, in which individuals indicate what level of noise they are willing to put up with while following speech, has been used to guide hearing aid fitting decisions and has been found to relate to prospective hearing aid use. Unlike objective measures of speech perception ability, ANL outcome is not related to individual hearing loss or age, but rather reflects an individual’s inherent acceptance of competing noise while listening to speech. As such, the measu...

  18. Type of speech material affects Acceptable Noise Level test outcome

    OpenAIRE

    Xaver eKoch; Gertjan eDingemanse; André eGoedegebure; Esther eJanse

    2016-01-01

    The Acceptable Noise Level (ANL) test, in which individuals indicate what level of noise they are willing to put up with while following speech, has been used to guide hearing aid fitting decisions and has been found to relate to prospective hearing aid use. Unlike objective measures of speech perception ability, ANL outcome is not related to individual hearing loss or age, but rather reflects an individual's inherent acceptance of competing noise while listening to speech. As such, the measu...

  19. ISOLOK VALVE ACCEPTANCE TESTING FOR DWPF SME SAMPLING PROCESS

    Energy Technology Data Exchange (ETDEWEB)

    Edwards, T.; Hera, K.; Coleman, C.; Jones, M.; Wiedenman, B.

    2011-12-05

    the two locations were compared to determine if the contents of the tank were well mixed. The Coliwasa sampler is a tube with a stopper at the bottom and is designed to obtain grab samples from specific locations within the drum contents. A position paper (4) was issued to address the prototypic flow loop issues and simulant selections. A statistically designed plan (5) was issued to address the total number of samples each sampler needed to pull, to provide the random order in which samples were pulled and to group samples for elemental analysis. The TTR required that the Isolok sampler perform as well as the Hydragard sampler during these tests to ensure the acceptability of the Isolok sampler for use in the DWPF sampling cells. Procedure No.L9.4-5015 was used to document the sample parameters and process steps. Completed procedures are located in R&D Engineering job folder 23269.

  20. 40 CFR 53.66 - Test procedure: Volatility test.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 5 2010-07-01 2010-07-01 false Test procedure: Volatility test. 53.66... Characteristics of Class II Equivalent Methods for PM2.5 § 53.66 Test procedure: Volatility test. (a) Overview. This test is designed to ensure that the candidate method's losses due to volatility when sampling...

  1. Lotka's Law: A Testing Procedure.

    Science.gov (United States)

    Pao, Miranda Lee

    1985-01-01

    A step-by-step outline is presented for testing the applicability of Lotka's law of scientific productivity. Steps include the computation of the values of the exponent and constant based on Lotka's method and the test for significance of observed frequency distribution against the estimated theoretical distribution derived from Lotka's formula.…

  2. Caustic addition system operability test procedure

    International Nuclear Information System (INIS)

    This test procedure provides instructions for performing operational testing of the major components of the 241-AN-107 Caustic Addition System by WHC and Kaiser personnel at the Rotating Equipment Shop run-in pit (Bldg. 272E)

  3. Standard-B auto grab sampler hydrogen monitoring system, Acceptance Test Report

    Energy Technology Data Exchange (ETDEWEB)

    Lott, D.T.

    1995-05-18

    Project W-369, Watch List Tank Hydrogen Monitors, installed a Standard-C Hydrogen Monitoring System (SHMS) on the Flammable gas waste tank AN-104. General Support Projects (8K510) was support by Test Engineering (7CH30) in the performance of the Acceptance Test Procedures (ATP) to qualify the SHMS cabinets on the waste tank. The ATP`s performance was controlled by Tank Farm work package. This completed ATP is transmitted by EDT-601748 as an Acceptance Test Report (ATR) in accordance with WHC-6-1, EP 4.2 and EP 1.12.

  4. Standard-B auto grab sampler hydrogen monitoring system, Acceptance Test Report

    International Nuclear Information System (INIS)

    Project W-369, Watch List Tank Hydrogen Monitors, installed a Standard-C Hydrogen Monitoring System (SHMS) on the Flammable gas waste tank AN-104. General Support Projects (8K510) was support by Test Engineering (7CH30) in the performance of the Acceptance Test Procedures (ATP) to qualify the SHMS cabinets on the waste tank. The ATP's performance was controlled by Tank Farm work package. This completed ATP is transmitted by EDT-601748 as an Acceptance Test Report (ATR) in accordance with WHC-6-1, EP 4.2 and EP 1.12

  5. Multiport riser and flange assemblies acceptance test report

    International Nuclear Information System (INIS)

    This document presents the results of the acceptance test for the multiport riser (MPR) and multiport flange (MPF) assemblies. The accepted MPR and MPF assemblies will be used in support of the hydrogen mitigation project for double-shell waste tank 241-SY-101 and other related projects. The testing described in this document verifies that the mechanical and interface features are operating as designed and that the unit is ready for field service. The objectives of the acceptance testing were as follows: Basic equipment functions and mechanical interfaces were verified; Installation and removal of equipment were demonstrated to the degree possible; Operation of the decon spray system and all valving was confirmed; and the accumulated leak rate of the MPR and MPF assemblies was determined

  6. Acceptance test for graphite components and construction status of HTTR

    International Nuclear Information System (INIS)

    In March, 1991, the Japan Atomic Energy Research Institute (JAERI) started to constructed the High Temperature engineering Test Reactor(HTTR) which is a 30-MW(thermal) helium gas-cooled reactor with a core composed of prismatic graphite blocks piled on the core support graphite structures. Two types of graphite materials are used in the HTTR. One is the garde IG-110, isotropic fine grain graphite, another is the grade PGX, medium-to-fine grained molded graphite. These materials were selected on the basis of the appropriate properties required by the HTTR reactor design. Industry-wide standards for an acceptance test of graphite materials used as main components of a nuclear reactor had not been established. The acceptance standard for graphite components of the HTTR, therefore, was drafted by JAERI and reviewed by specialists outside JAERI. The acceptance standard consists of the material testing, non-destructive examination such as the ultrasonic and eddy current testings, dimensional and visual inspections and assembly test. Ultrasonic and eddy current testings are applied to graphite logs to detect an internal flaw and to graphite components to detect a surface flaw, respectively. The assembly test is performed at the works, prior to their installation in the reactor pressure vessel, to examine fabricating precision of each component and alignment of piled-up structures. The graphite components of the HTTR had been tested on the basis of the acceptance standard. It was confirmed that the graphite manufacturing process was well controlled and high quality graphite components were provided to the HTTR. All graphite components except for the fuel graphite blocks are to be installed in the reactor pressure vessel of the HTTR in September 1995. The paper describes the construction status of the HTTR focusing on the graphite components. The acceptance test results are also presented in this paper. (author). Figs

  7. 40 CFR 53.65 - Test procedure: Loading test.

    Science.gov (United States)

    2010-07-01

    ... operating manual referred to in section 7.4.18 of 40 CFR part 50, appendix L. (3) Determine the preweight of... 40 Protection of Environment 5 2010-07-01 2010-07-01 false Test procedure: Loading test. 53.65... Characteristics of Class II Equivalent Methods for PM2.5 § 53.65 Test procedure: Loading test. (a) Overview....

  8. W-026 acceptance test report system integration equipment (SIE)(submittal {number_sign} 018.6.A)

    Energy Technology Data Exchange (ETDEWEB)

    Watson, T.L.

    1997-01-27

    Acceptance testing of the System Integration Equipment (SIE) at Hanford was performed in two stages. The first was inconclusive, and resulted in a number of findings. These finding. are summarized as part of this report. The second stage of testing addressed these findings, and performed full system testing per the approved test procedure. This report includes summaries of all testing, results and finding.. Although the SIE did not in some cases perform as required for plant operations, it did perform per the system specification. (These discrepancies were noted and are addressed elsewhere.) Following testing, the system was formaLLy accepted. Documentation of this acceptance is incLuded in this report.

  9. NEVADA TEST SITE WASTE ACCEPTANCE CRITERIA, JUNE 2006

    Energy Technology Data Exchange (ETDEWEB)

    U.S. DEPARTMENT OF ENERGY, NATIONAL NUCLEAR SECURITY ADMINISTRATION NEVADA SITE OFFICE

    2006-06-01

    This document establishes the U.S. Department of Energy (DOE), National Nuclear Security Administration Nevada Site Office (NNSA/NSO) waste acceptance criteria (WAC). The WAC provides the requirements, terms, and conditions under which the Nevada Test Site (NTS) will accept low-level radioactive (LLW) and mixed waste (MW) for disposal. It includes requirements for the generator waste certification program, characterization, traceability, waste form, packaging, and transfer. The criteria apply to radioactive waste received at the NTS Area 3 and Area 5 Radioactive Waste Management Complex (RWMC) for storage or disposal.

  10. W-026, acceptance test report TRU empty drum compactor (submittal{number_sign}634)

    Energy Technology Data Exchange (ETDEWEB)

    Watson, T.L.

    1997-06-18

    On 10/19/96 and 10/21/96, the 06/04/96 version of the Empty Drum Compactor Acceptance Test Procedure was used to perform tests by INET Corporation at the Hanford WRAP facility. The INET compaction components were installed in the Diversified glovebox. The Diversified glovebox and cart and the INET compaction components, EDC HPU and drum centering device constitute the Empty Drum Compactor.

  11. Startup of the FFTF sodium cooled reactor. [Acceptance Test Program

    Energy Technology Data Exchange (ETDEWEB)

    Redekopp, R.D.; Umek, A.M.

    1981-03-01

    The Fast Flux Test Facility (FFTF), located on the Department of Energy (DOE) Hanford Reservation near Richland, Washington, is a 3 Loop 400 MW(t) sodium cooled fast reactor with a primary mission to test fuels and materials for development of the Liquid Metal Fast Breeder Reactor (LMFBR). Bringing FFTF to a condition to accomplish this mission is the goal of the Acceptance Test Program (ATP). This program was the mechanism for achieving startup of the FFTF. Highlights of the ATP involving the system inerting, liquid metal and inerted cell testing and initial ascent to full power are discussed.

  12. TMACS test procedure TP005: Sensor configuration, logging, and data conversion. Revision 4

    International Nuclear Information System (INIS)

    The TMACS Software Project Test Procedures translate the project's acceptance criteria into test steps. Software releases are certified when the affected Test Procedures are successfully performed and the customers authorize installation of these changes. This Test Procedure addresses the sensor configuration, conversion and logging requirements of the TMACS. The features to be tested are as follows: sensor configuration data; conversion of continuous sensor data to engineering units; conversion of digital data to discrete states; discrete sensor data logging; and continuous sensor data logging

  13. Acceptance of genetic testing in a general population

    DEFF Research Database (Denmark)

    Aro, A R; Hakonen, A; Hietala, M;

    1997-01-01

    The aim of the study was to analyze effects of age, education and gender on acceptance of genetic testing. Subjects, n = 1967 aged 15-69, were a stratified random sample of the Finnish population. One thousand, one hundred and sixty nine subjects, 530 men and 639 women, returned the questionnaire....... The majority of the respondents approved of the availability of genetic testing. Young, aged 15-24, were more favourable towards testing and more willing to undergo suggested tests, but they were also more worried than others about the misuse of test results. Men aged 45-69 with only basic education were more...... in favour of mandatory genetic testing than other respondents. Respondents with university education were more critical towards genetic testing and expressed their worry about eugenics more often than other education groups. In conclusion, there are age, education and gender related differences...

  14. Central Acceptance Testing for Camera Technologies for CTA

    CERN Document Server

    Bonardi, A; Chadwick, P; Dazzi, F; Förster, A; Hörandel, J R; Punch, M

    2015-01-01

    The Cherenkov Telescope Array (CTA) is an international initiative to build the next generation ground based very-high energy gamma-ray observatory. It will consist of telescopes of three different sizes, employing several different technologies for the cameras that detect the Cherenkov light from the observed air showers. In order to ensure the compliance of each camera technology with CTA requirements, CTA will perform central acceptance testing of each camera technology. To assist with this, the Camera Test Facilities (CTF) work package is developing a detailed test program covering the most important performance, stability, and durability requirements, including setting up the necessary equipment. Performance testing will include a wide range of tests like signal amplitude, time resolution, dead-time determination, trigger efficiency, performance testing under temperature and humidity variations and several others. These tests can be performed on fully-integrated cameras using a portable setup at the came...

  15. Surface moisture measurement system hardware acceptance test report

    Energy Technology Data Exchange (ETDEWEB)

    Ritter, G.A., Westinghouse Hanford

    1996-05-28

    This document summarizes the results of the hardware acceptance test for the Surface Moisture Measurement System (SMMS). This test verified that the mechanical and electrical features of the SMMS functioned as designed and that the unit is ready for field service. The bulk of hardware testing was performed at the 306E Facility in the 300 Area and the Fuels and Materials Examination Facility in the 400 Area. The SMMS was developed primarily in support of Tank Waste Remediation System (TWRS) Safety Programs for moisture measurement in organic and ferrocyanide watch list tanks.

  16. Test procedure for boxed waste assay system

    International Nuclear Information System (INIS)

    This document, prepared by Los Alamos National Laboratory's NMT-4 group, details the test methodology and requirements for Acceptance/Qualification testing of a Boxed Waste Assay System (BWAS) designed and constructed by Pajarito Scientific Corporation. Testing of the BWAS at the Plutonium Facility (TA55) at Los Alamos National Laboratory will be performed to ascertain system adherence to procurement specification requirements. The test program shall include demonstration of conveyor handling capabilities, gamma ray energy analysis, and imaging passive/active neutron accuracy and sensitivity. Integral to these functions is the system's embedded operating and data reduction software

  17. MCO combustible gas management leak test acceptance criteria

    Energy Technology Data Exchange (ETDEWEB)

    SHERRELL, D.L.

    1999-05-11

    Existing leak test acceptance criteria for mechanically sealed and weld sealed multi-canister overpacks (MCO) were evaluated to ensure that MCOs can be handled and stored in stagnant air without compromising the Spent Nuclear Fuel Project's overall strategy to prevent accumulation of combustible gas mixtures within MCO's or within their surroundings. The document concludes that the integrated leak test acceptance criteria for mechanically sealed and weld sealed MCOs (1 x 10{sup -5} std cc/sec and 1 x 10{sup -7} std cc/sec, respectively) are adequate to meet all current and foreseeable needs of the project, including capability to demonstrate compliance with the NFPA 60 Paragraph 3-3 requirement to maintain hydrogen concentrations [within the air atmosphere CSB tubes] t or below 1 vol% (i.e., at or below 25% of the LFL).

  18. Central Acceptance Testing for Camera Technologies for CTA

    OpenAIRE

    Bonardi, A.; T. Buanes; Chadwick, P.; Dazzi, F.; A. Förster(CERN, Geneva, Switzerland); Hörandel, J. R.; Punch, M.; Consortium, R. M. Wagner for the CTA

    2015-01-01

    The Cherenkov Telescope Array (CTA) is an international initiative to build the next generation ground based very-high energy gamma-ray observatory. It will consist of telescopes of three different sizes, employing several different technologies for the cameras that detect the Cherenkov light from the observed air showers. In order to ensure the compliance of each camera technology with CTA requirements, CTA will perform central acceptance testing of each camera technology. To assist with thi...

  19. Designing user acceptance testing plan for ERP implementation

    OpenAIRE

    Lietzen, Ari

    2013-01-01

    The user acceptance testing (UAT) is one of the success factors in ERP (Enterprise Resource Planning) implementation. ERP systems can be very complicated with integrated business processes. ERP software is one of the most important business applications in the company. Companies carry out business processes with ERP. One of the critical business applications to the companies is ERP software. Many ERP implementation projects have failed during the past years. One of the most common reasons is ...

  20. Planning and acceptance testing of Mv therapy installations

    International Nuclear Information System (INIS)

    Purpose/Objective: The purpose of this course is to ensure that the individuals responsible for acquiring megavoltage therapy equipment are aware of all aspects involved. The object is to cover in a broad overview the considerations that go into selecting, installing, testing and accepting megavoltage therapy equipment and to provide a resource for more detailed information. Planning and acceptance testing of a megavoltage therapy installation is a major undertaking for any size group, institution or department. It can take approximately two years from the time a decision is made to get a machine to where the first patient is treated. Because the equipment and site preparation are expensive, and most of the machines are complex and can be supplied by several different manufacturers, it is imperative that a great deal of thought and care go into the decision: the aim being that the specifications of the machine that is installed meet the immediate need of the department and the needs for the projected lifetime of the equipment. A general survey of the types of equipment available will be presented. This will concentrate on general purpose linear accelerators, although Cobalt 60 machines, microtrons and special purpose machines (intra-operative equipment) will also be covered. General descriptions of the machines along with typical specifications will be given. Selecting the best machine to meet specific needs can be quite complex and criteria for making the selection are presented in a series of twelve steps. Site selection and room design, including the console area, are also critical. The general principles for shielding calculations will be provided. Critical to any installation is the acceptance testing of the equipment including material and radiation oncology performance tests. These tests will be outlined. A reading list of suitable references describing in detail many of the aspects of this course will be provided

  1. Planning and acceptance testing of MV therapy installations

    International Nuclear Information System (INIS)

    Purpose/Objective: The purpose of this course is to ensure that the individuals responsible for acquiring megavoltage therapy equipment are aware of all aspects involved. The object is to cover in a broad overview the considerations that go into selecting, installing, testing and accepting megavoltage therapy equipment and to provide a resource for more detailed information. Planning and acceptance testing of a megavoltage therapy installation is a major undertaking for any size group, institution or department. It can take approximately two years from the time a decision is made to get a machine to where the first patient is treated. Because the equipment and site preparation are expensive, and most of the machines are complex and can be supplied by several different manufacturers, it is imperative that a great deal of thought and care go into the decision: the aim being that the specifications of the machine that is installed meet the immediate need of the department and the needs for the projected lifetime of the equipment. A general survey of the types of equipment available will be presented. This will concentrate on general purpose linear accelerators, although Cobalt 60 machines, microtrons and special purpose machines (intra-operative equipment) will also be covered. General descriptions of the machines along with typical specifications will be given. Selecting the best machine to meet specific needs can be quite complex and criteria for making the selection are presented in a series of twelve steps. Site selection and room design, including the console area, are also critical. The general principles for shielding calculations will be provided. Critical to any installation is the acceptance testing of the equipment including material and radiation oncology performance tests. These tests will be outlined. A reading list of suitable references describing in detail many of the aspects of this course will be provided

  2. Planning and acceptance testing of MV therapy installations

    International Nuclear Information System (INIS)

    Purpose: This course is designed for practitioners and beginners in brachytherapy. The aim is to review biological principles equipment are aware of all aspects involved. The object is to cover in a broad overview the considerations that go into selecting, installing, testing and accepting megavoltage therapy equipment and to provide a resource for more detailed information. Planning and acceptance testing of a megavoltage therapy installation is a major undertaking for any size group, institution or department. It can take approximately two years from the time a decision is made to get a machine to where the first patient is treated. Because the equipment and site preparation are expensive, and most of the machines are complex and can be supplied by several different manufacturers, it is imperative that a great deal of thought and care go into the decision: the aim being that the specifications of the machine that is installed meet the immediate need of the department and the needs for the projected lifetime of the equipment. A general survey of the types of equipment available will be presented. This will concentrate on general purpose linear accelerators, although Cobalt 60 machines, microtrons and special purpose machines (intra-operative equipment) will also be covered. General descriptions of the machines along with typical specifications will be given. Selecting the best machine to meet specific needs can be quite complex and criteria for making the selection are presented in a series of twelve steps. site selection and room design, including the console area, are also critical. The general principles for shielding calculations will be provided. Critical to any installation is the acceptance testing of the equipment including material and radiation oncology performance tests. These tests will be outlined. A reading list of suitable references describing in detail many of the aspects of this course will be provided

  3. Nevada Test Site waste acceptance criteria [Revision 1

    International Nuclear Information System (INIS)

    Revision one updates the requirements, terms, and conditions under which the Nevada Test Site (NTS) will accept low-level radioactive and mixed waste for disposal; and transuranic and transuranic mixed waste for interim storage at the NTS. Review each section of this document. This document is not intended to include all of the requirements; rather, it is meant as a guide toward meeting the regulations. All references in this document should be observed to avoid omission of requirements on which acceptance or rejection of waste will be based. The Department of Energy/Nevada Operations Office (DOE/NV) and support contractors are available to assist you in understanding or interpreting this document

  4. Nevada Test Site waste acceptance criteria [Revision 1

    Energy Technology Data Exchange (ETDEWEB)

    None

    1997-08-01

    Revision one updates the requirements, terms, and conditions under which the Nevada Test Site (NTS) will accept low-level radioactive and mixed waste for disposal; and transuranic and transuranic mixed waste for interim storage at the NTS. Review each section of this document. This document is not intended to include all of the requirements; rather, it is meant as a guide toward meeting the regulations. All references in this document should be observed to avoid omission of requirements on which acceptance or rejection of waste will be based. The Department of Energy/Nevada Operations Office (DOE/NV) and support contractors are available to assist you in understanding or interpreting this document.

  5. Acceptance Test Report for Fourth-Generation Hanford Corrosion Monitoring Cabinet

    International Nuclear Information System (INIS)

    This Acceptance Test Plan (ATP) will document the satisfactory operation of the third-generation corrosion monitoring cabinet (Hiline Engineering Part No.0004-CHM-072-C01). This ATP will be performed by the manufacturer of the cabinet prior to delivery to the site. The objective of this procedure is to demonstrate and document the acceptance of the corrosion monitoring cabinet. The test will consist of a continuity test of the cabinet wiring from the end of cable to be connected to corrosion probe, through the appropriate intrinsic safety barriers and out to the 15 pin D-shell connectors to be connected to the corrosion monitoring instrument. Additional testing will be performed using a constant current and voltage source provided by the corrosion monitoring hardware manufacturer to verify proper operation of corrosion monitoring instrumentation

  6. Trust, confidence, procedural fairness, outcome fairness, moral conviction, and the acceptance of GM field experiments.

    Science.gov (United States)

    Siegrist, Michael; Connor, Melanie; Keller, Carmen

    2012-08-01

    In 2005, Swiss citizens endorsed a moratorium on gene technology, resulting in the prohibition of the commercial cultivation of genetically modified crops and the growth of genetically modified animals until 2013. However, scientific research was not affected by this moratorium, and in 2008, GMO field experiments were conducted that allowed us to examine the factors that influence their acceptance by the public. In this study, trust and confidence items were analyzed using principal component analysis. The analysis revealed the following three factors: "economy/health and environment" (value similarity based trust), "trust and honesty of industry and scientists" (value similarity based trust), and "competence" (confidence). The results of a regression analysis showed that all the three factors significantly influenced the acceptance of GM field experiments. Furthermore, risk communication scholars have suggested that fairness also plays an important role in the acceptance of environmental hazards. We, therefore, included measures for outcome fairness and procedural fairness in our model. However, the impact of fairness may be moderated by moral conviction. That is, fairness may be significant for people for whom GMO is not an important issue, but not for people for whom GMO is an important issue. The regression analysis showed that, in addition to the trust and confidence factors, moral conviction, outcome fairness, and procedural fairness were significant predictors. The results suggest that the influence of procedural fairness is even stronger for persons having high moral convictions compared with persons having low moral convictions. PMID:22150405

  7. Parametric Sensitivity Tests- European PEM Fuel Cell Stack Test Procedures

    DEFF Research Database (Denmark)

    Araya, Samuel Simon; Andreasen, Søren Juhl; Kær, Søren Knudsen

    2014-01-01

    As fuel cells are increasingly commercialized for various applications, harmonized and industry-relevant test procedures are necessary to benchmark tests and to ensure comparability of stack performance results from different parties. This paper reports the results of parametric sensitivity tests...

  8. Waste retrieval sluicing system data acquisition system acceptance test report

    International Nuclear Information System (INIS)

    This document describes the test procedure for the Project W-320 Tank C-106 Sluicing Data Acquisition System (W-320 DAS). The Software Test portion will test items identified in the WRSS DAS System Description (SD), HNF-2115. Traceability to HNF-2115 will be via a reference that follows in parenthesis, after the test section title. The Field Test portion will test sensor operability, analog to digital conversion, and alarm setpoints for field instrumentation. The W-320 DAS supplies data to assist thermal modeling of tanks 241-C-106 and 241-AY-102. It is designed to be a central repository for information from sources that would otherwise have to be read, recorded, and integrated manually. Thus, completion of the DAS requires communication with several different data collection devices and output to a usable PC data formats. This test procedure will demonstrate that the DAS functions as required by the project requirements stated in Section 3 of the W-320 DAS System Description, HNF-2115

  9. Could quantitative liver function tests gain wide acceptance among hepatologists?

    Institute of Scientific and Technical Information of China (English)

    Giovanni Tarantino

    2009-01-01

    It has been emphasized that the assessment of residual liver function is of paramount importance to determine the following: severity of acute or chronic liver diseases independent of etiology; long-term prognosis; step-bystep disease progression; surgical risk; and efficacy of antiviral treatment. The most frequently used tools are the galactose elimination capacity to asses hepatocyte cytosol activity, plasma clearance of indocyanine green to assess excretory function, and antipyrine clearance to estimate microsomal activity. However, a widely accepted liver test (not necessarily a laboratory one) to assess quantitative functional hepatic reserve still needs to be established, although there have been various proposals. Furthermore, who are the operators that should order these tests? Advances in analytic methods are expected to allow quantitative liver function tests to be used in clinical practice.

  10. [Clinical trial data validation and user acceptance testing].

    Science.gov (United States)

    Sun, Hua-long; Dai, Nan

    2015-11-01

    For pharmaceutical industries, clinical data is one of the most valuable deliverables. It is also the basis of analysis, submission, approval, labeling and marketing of a drug product. To ensure the integrity and reliability of clinical data, a scientific standardized quality control (QC) has to be established at each step of a clinical trial. Data validation is conducted to ensure the reasonability and compliance of clinical data by checking data quality before the data is statistically analyzed. This paper focuses on purpose of data validation, creation of data validation plan, rationale of data validation, types of data validation and performance of user acceptance testing on clinical database. PMID:26911047

  11. The Cloze Procedure as a Progress Test.

    Science.gov (United States)

    Stansfield, Charles

    This paper reports on a pilot study conducted to determine the possible use of the cloze procedure as a substitute measure of achievement in a second-year Spanish culture and civilization class. Twenty students enrolled in a third-semester Spanish class at the University of Colorado were simultaneously given multiple choice tests and cloze tests…

  12. Acceptance testing and quality assurance of Simulix evolution radiotherapy simulator

    International Nuclear Information System (INIS)

    The success of radiotherapy depends on precise treatment simulation and proper patient positioning. The simulator is a conventional radiographic and fluoroscopic system which emulates the geometrical positions of radiotherapy treatment unit. Hence, the acceptance tests and quality assurance (QA) of the simulator are important prior to its commissioning for the safe and precise clinical use. The verification of mechanical and optical readouts, field size, isocenter, optical and radiation field congruence were performed. The X-ray beam parameters were tested for kVp, mAs and consistency of radiation output. The flat panel detector performance was checked with respect to resolution, low contrast sensitivity (LCS), automatic dose rate control (ADRC), and gray image resolution (GIR). Gantry, table, and imaging system collision possibility was checked. Radiation survey around the room was also performed. The field size test for digital readout and on graph paper, the results of isocenter checkup for rotation of gantry, collimator, and couch, and the deviations observed in auto stop for various movements were found within the tolerance limits. Optical field and radiation field was found congruent. All the lasers were found aligned with the established isocenter. Maximum deviation for set and measured kV was found to be 3% in fluoro mode. The maximum deviation observed in mAs was 1.5% in 3-point as well as in 2-point film exposed mode. The X-ray output was found consistent. The results of tests for resolution, LCS, ADRC, and GIR of the flat panel detector were within tolerance limits. All the six safety interlocks were found working. Radiation level around the room was found within the acceptable limits. All the tests carried out were found within the tolerance limits. The data which has been taken in this study will provide basic support to the routine QA of the simulator. (author)

  13. Topographical mapping system for radiological and hazardous environments acceptance testing

    International Nuclear Information System (INIS)

    During the summer of 1996, the Topographical Mapping System (TMS) for hazardous and radiological environments and its accompanying three-dimensional (3-D) visualization tool, the Interactive Computer-Enhanced Remote-Viewing System (ICERVS), were delivered to Oak Ridge National Laboratory (ORNL). ORNL and Mechanical Technology, Inc., performed final acceptance testing of the TMS during the next eight months. The TMS was calibrated and characterized during this period. This paper covers the calibration, characterization, and acceptance testing of the TMS. Development of the TMS and ICERVS was initiated by the US Department of Energy (DOE) for the purpose of characterization and remediation of underground storage tanks (USTs) at DOE sites across the country. DOE required a 3-D, topographical mapping system suitable for use in hazardous and radiological environments. The intended application is the mapping of the interior of USTs as part of DOE's waste characterization and remediation efforts and to obtain baseline data on the content of the storage tank interiors as well as data on changes in the tank contents and levels brought about by waste remediation steps. Initially targeted for deployment at the Hanford Washington site, the TMS is designed to be a self-contained, compact, and reconfigurable system that is capable of providing rapid, variable-resolution mapping information in poorly characterized workspaces with a minimum of operator intervention

  14. Topographical mapping system for radiological and hazardous environments acceptance testing

    Science.gov (United States)

    Armstrong, Gary A.; Dochat, G. R.

    1997-09-01

    During the summer of 1996, the topographical mapping system (TMS) for hazardous and radiological environments and its accompanying three-dimensional (3-D) visualization tool, the interactive computer-enhanced remote-viewing system (ICERVS), were delivered to Oak Ridge National Laboratory (ORNL). ORNL and Mechanical Technology, Inc., performed final acceptance testing of the TMS during the next eight months. The TMS was calibrated and characterized during this period. This paper covers the calibration, characterization, and acceptance testing of the TMS. Development of the TMS and the ICERVS was initiated by the U.S. Department of Energy (DOE) for the purpose of characterization and remediation of underground storage tanks (USTs) at DOE sites across the country. DOE required a 3-D, topographical mapping system suitable for use in hazardous and radiological environments. The intended application is the mapping of the interior of USTs as part of DOE's waste characterization and remediation efforts and to obtain baseline data on the content of the storage tank interiors as well as data on changes in the tank contents and levels brought about by waste remediation steps. Initially targeted for deployment at the Hanford Washington site, the TMS is designed to be a self-contained, compact, reconfigurable system that is capable of providing rapid, variable-resolution mapping information in poorly characterized workspaces with a minimum of operator intervention.

  15. Type of speech material affects Acceptable Noise Level test outcome

    Directory of Open Access Journals (Sweden)

    Xaver eKoch

    2016-02-01

    Full Text Available The Acceptable Noise Level (ANL test, in which individuals indicate what level of noise they are willing to put up with while following speech, has been used to guide hearing aid fitting decisions and has been found to relate to prospective hearing aid use. Unlike objective measures of speech perception ability, ANL outcome is not related to individual hearing loss or age, but rather reflects an individual's inherent acceptance of competing noise while listening to speech. As such, the measure may predict aspects of hearing aid success. Crucially, however, recent studies have questioned its repeatability (test-retest reliability. The first question for this study was whether the inconsistent results regarding the repeatability of the ANL test may be due to differences in speech material types used in previous studies. Second, it is unclear whether meaningfulness and semantic coherence of the speech modify ANL outcome. To investigate these questions, we compared ANLs obtained with three types of materials: the International Speech Test Signal (ISTS, which is non-meaningful and semantically non-coherent by definition, passages consisting of concatenated meaningful standard audiology sentences, and longer fragments taken from conversational speech. We included conversational speech as this type of speech material is most representative of everyday listening. Additionally, we investigated whether ANL outcomes, obtained with these three different speech materials, were associated with self-reported limitations due to hearing problems and listening effort in everyday life, as assessed by a questionnaire. ANL data were collected for 57 relatively good-hearing adult participants with an age range representative for hearing aid users. Results showed that meaningfulness, but not semantic coherence of the speech material affected ANL. Less noise was accepted for the non-meaningful ISTS signal than for the meaningful speech materials. ANL repeatability was

  16. Type of Speech Material Affects Acceptable Noise Level Test Outcome.

    Science.gov (United States)

    Koch, Xaver; Dingemanse, Gertjan; Goedegebure, André; Janse, Esther

    2016-01-01

    The acceptable noise level (ANL) test, in which individuals indicate what level of noise they are willing to put up with while following speech, has been used to guide hearing aid fitting decisions and has been found to relate to prospective hearing aid use. Unlike objective measures of speech perception ability, ANL outcome is not related to individual hearing loss or age, but rather reflects an individual's inherent acceptance of competing noise while listening to speech. As such, the measure may predict aspects of hearing aid success. Crucially, however, recent studies have questioned its repeatability (test-retest reliability). The first question for this study was whether the inconsistent results regarding the repeatability of the ANL test may be due to differences in speech material types used in previous studies. Second, it is unclear whether meaningfulness and semantic coherence of the speech modify ANL outcome. To investigate these questions, we compared ANLs obtained with three types of materials: the International Speech Test Signal (ISTS), which is non-meaningful and semantically non-coherent by definition, passages consisting of concatenated meaningful standard audiology sentences, and longer fragments taken from conversational speech. We included conversational speech as this type of speech material is most representative of everyday listening. Additionally, we investigated whether ANL outcomes, obtained with these three different speech materials, were associated with self-reported limitations due to hearing problems and listening effort in everyday life, as assessed by a questionnaire. ANL data were collected for 57 relatively good-hearing adult participants with an age range representative for hearing aid users. Results showed that meaningfulness, but not semantic coherence of the speech material affected ANL. Less noise was accepted for the non-meaningful ISTS signal than for the meaningful speech materials. ANL repeatability was comparable across

  17. 40 CFR 53.64 - Test procedure: Static fractionator test.

    Science.gov (United States)

    2010-07-01

    ... sampler's operating manual referred to in section 7.4.18 of 40 CFR part 50, appendix L. (2) Generate... 40 Protection of Environment 5 2010-07-01 2010-07-01 false Test procedure: Static fractionator test. 53.64 Section 53.64 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED)...

  18. General Mission Analysis Tool (GMAT) Acceptance Test Plan [Draft

    Science.gov (United States)

    Dove, Edwin; Hughes, Steve

    2007-01-01

    The information presented in this Acceptance Test Plan document shows the current status of the General Mission Analysis Tool (GMAT). GMAT is a software system developed by NASA Goddard Space Flight Center (GSFC) in collaboration with the private sector. The GMAT development team continuously performs acceptance tests in order to verify that the software continues to operate properly after updates are made. The GMAT Development team consists of NASA/GSFC Code 583 software developers, NASA/GSFC Code 595 analysts, and contractors of varying professions. GMAT was developed to provide a development approach that maintains involvement from the private sector and academia, encourages collaborative funding from multiple government agencies and the private sector, and promotes the transfer of technology from government funded research to the private sector. GMAT contains many capabilities, such as integrated formation flying modeling and MATLAB compatibility. The propagation capabilities in GMAT allow for fully coupled dynamics modeling of multiple spacecraft, in any flight regime. Other capabilities in GMAT inclucle: user definable coordinate systems, 3-D graphics in any coordinate system GMAT can calculate, 2-D plots, branch commands, solvers, optimizers, GMAT functions, planetary ephemeris sources including DE405, DE200, SLP and analytic models, script events, impulsive and finite maneuver models, and many more. GMAT runs on Windows, Mac, and Linux platforms. Both the Graphical User Interface (GUI) and the GMAT engine were built and tested on all of the mentioned platforms. GMAT was designed for intuitive use from both the GUI and with an importable script language similar to that of MATLAB.

  19. Acceptance tests of cryogenic components for SST-1

    International Nuclear Information System (INIS)

    The cryogenics system is one of the critical sub-systems in SST-1. The role of cryogenics systems is to facilitate the smooth cool-down, warm up and steady state operation as well as to take care about any abnormal events in safe mode during the operation. As a part of SST-1 machine shell, the cryogenics division is responsible for cold helium (at 4.5 K) and liquid nitrogen (at 77 K) hydraulics. This hydraulic network consists of headers, 80 K embossed shields, interconnecting loops, ceramic feed through at the vacuum barrier, flexible hoses and electrical breaks for cold helium as well as liquid nitrogen circuits. Before installation of individual components in SST-1, the performance validation tests were mandatory as part of mandatory stringent protocols fulfillment. The performance validation tests were conducted from individual components to integrated assembly level. In this paper, we report the essence of procedures followed and the performance test results. (author)

  20. Test marketing and consumer acceptance of irradiated meat products

    International Nuclear Information System (INIS)

    This study consists of two parts: irradiation processing of cooked meat and irradiation preservation of prepackaged chilled fresh cut meats. Irradiation of prepackaged pickled meat products dipped in grains stillage at a dose 6-8 kGy eliminated common food-borne microorganisms, such as E. Coli and other microbial pathogens and extended the shelf life of the product to 10 days at 5 deg. C. Test marketing of 40,000 bags (about 10,000 kg) of the product in more than 100 supermarkets in the city of Shanghai showed no untoward problem with consumer acceptance. Irradiation of prepackaged chilled fresh cut pork at a dose 3 kGy led to inactivation of microbial pathogens and parasites with a concomitant reduction in numbers of common spoilage microorganisms and extension of shelf life of the product for 30 days at 5 deg. C. The cost benefit and marketing applications were evaluated. (author)

  1. 118-B-1 excavation treatability test procedures

    International Nuclear Information System (INIS)

    This treatability study has two purposes: to support development of the approach to be used for burial ground remediation, and to provide specific engineering information for the design of burial grounds receiving waste generated from the 100 Area removal actions. Data generated from this test will also provide performance and cost information necessary for detailed analysis of alternatives for burial ground remediation. Further details on the test requirements, milestones and data quality objectives are described in detail in the 118-B-1 Excavation Treatability Test Plan (DOE/RL-94-43). These working procedures are intended for use by field personnel to implement the requirements of the milestone. A copy of the detailed Test Plan will be kept on file at the on-site field support trailer, and will be available for review by field personnel

  2. 118-B-1 excavation treatability test procedures

    Energy Technology Data Exchange (ETDEWEB)

    Frain, J.M.

    1994-08-01

    This treatability study has two purposes: to support development of the approach to be used for burial ground remediation, and to provide specific engineering information for the design of burial grounds receiving waste generated from the 100 Area removal actions. Data generated from this test will also provide performance and cost information necessary for detailed analysis of alternatives for burial ground remediation. Further details on the test requirements, milestones and data quality objectives are described in detail in the 118-B-1 Excavation Treatability Test Plan (DOE/RL-94-43). These working procedures are intended for use by field personnel to implement the requirements of the milestone. A copy of the detailed Test Plan will be kept on file at the on-site field support trailer, and will be available for review by field personnel.

  3. Acceptance test of graphite components in nuclear reactor

    International Nuclear Information System (INIS)

    The HTTR is the first high temperature gas-cooled reactor in Japan. It is a test reactor with thermal power of 30 MW and coolant outlet temperature of 950degC at maximum. To achieve high temperature coolant core internals were made of graphite and carbon materials due to their excellent thermal resistivity. After fabrication of graphite and carbon components at works they were installed in the HTTR, and now it is in the power up testing stage. Concerning the inspection standard of the graphite and carbon components, nondomestic standard exists as main components in the nuclear reactor. It is necessary, therefore, to prescribe the inspection standards for the HTTR graphite components. Many research and developments in relation to the inspection standard, e.g. in the research field of nondestructive examination of the graphite material, had been performed, and then the JAERI established the inspection standard. The acceptance test of the graphite and carbon components was carried out based on the inspection standard. This paper prescribes the outline of the established inspection standard. (author)

  4. Chemical Compatibility Testing Final Report Including Test Plans and Procedures

    International Nuclear Information System (INIS)

    This report provides an independent assessment of information on mixed waste streams, chemical compatibility information on polymers, and standard test methods for polymer properties. It includes a technology review of mixed low-level waste (LLW) streams and material compatibilities, validation for the plan to test the compatibility of simulated mixed wastes with potential seal and liner materials, and the test plan itself. Potential packaging materials were reviewed and evaluated for compatibility with expected hazardous wastes. The chemical and physical property measurements required for testing container materials were determined. Test methodologies for evaluating compatibility were collected and reviewed for applicability. A test plan to meet US Department of Energy and Environmental Protection Agency requirements was developed. The expected wastes were compared with the chemical resistances of polymers, the top-ranking polymers were selected for testing, and the most applicable test methods for candidate seal and liner materials were determined. Five recommended solutions to simulate mixed LLW streams are described. The test plan includes descriptions of test materials, test procedures, data collection protocols, safety and environmental considerations, and quality assurance procedures. The recommended order of testing to be conducted is specified

  5. Solar panel acceptance testing using a pulsed solar simulator

    Science.gov (United States)

    Hershey, T. L.

    1977-01-01

    Utilizing specific parameters as area of an individual cell, number in series and parallel, and established coefficient of current and voltage temperature dependence, a solar array irradiated with one solar constant at AMO and at ambient temperature can be characterized by a current-voltage curve for different intensities, temperatures, and even different configurations. Calibration techniques include: uniformity in area, depth and time, absolute and transfer irradiance standards, dynamic and functional check out procedures. Typical data are given for individual cell (2x2 cm) to complete flat solar array (5x5 feet) with 2660 cells and on cylindrical test items with up to 10,000 cells. The time and energy saving of such testing techniques are emphasized.

  6. Procedure guideline for radioiodine test. Version

    International Nuclear Information System (INIS)

    The version 2 of the procedure guideline for radioiodine test is an update of the guideline published in 1999. The following statements were added or modified: The procedure guideline discusses the pros and cons of a single measurement or of repeated measurements of the iodine-131 uptake and their optimal timing. Different formulas are described when one, two or three values of the radioiodine kinetic are available. The probe with a sodium iodide crystal, alternative or additionally the gamma-camera using the ROI-technique are instrumentations for the measurement of iodine-131 uptake. A possible source of error is an inappropriate measurement (sonography) of the target volume. The patients' preparation includes the withdrawal of antithyroid drugs 2-3 days before radioiodine administration. The patient has to avoid iodine-containing medication and the possibility of additives of iodide in vitamin- or electrolyte-supplementation has to be considered. (orig.)

  7. Overcoming Denial through the Group: A Test of Acceptance Theory.

    Science.gov (United States)

    Rugel, Robert P.; Barry, Denise

    1990-01-01

    Found participants (N=28) in alcohol treatment groups showed decreases in denial of drinking problems and decreases in psychopathology following 12 weeks of group counseling. Determined greater self-acceptance was associated with experiencing acceptance by group and with greater decreases in denial; decreases in denial and psychopathology were…

  8. Specification and acceptance testing of nuclear medicine equipment

    International Nuclear Information System (INIS)

    The purchase of nuclear medicine equipment is of prime importance in the operation of a clinical service. Failure to properly evaluate the potential uses of the instrumentation and the various operational characteristics of the equipment can often result in the purchase of inappropriate or inferior instruments. The magnitude of the purchase in terms of time and financial investments make it imperative that the purchase be approached in a systematic manner. Consideration of both the intended clinical functions and personnel requirements is important. It is necessary also to evaluate the ability of the equipment vendor to support the instrumentation after the purchase has been completed and the equipment installed in the clinical site. The desired specifications of the instrument characteristics should be stated in terms that can be verified by acceptance testing. The complexity of modern instrumentation and the sensitivity of it to the environment require the buyer to take into account the potential problems of controlling the temperature, humidity, and electrical power of the installation site. If properly and systematically approached, the purchase of new nuclear medicine instrumentation can result in the acquisition of a powerful diagnostic tool which will have a useful lifetime of many years. If not so approached, it may result in the expenditure of a large amount of money and personnel time without the concomitant return in useful clinical service. (author)

  9. In situ vitrification large-scale operational acceptance test analysis

    International Nuclear Information System (INIS)

    A thermal treatment process is currently under study to provide possible enhancement of in-place stabilization of transuranic and chemically contaminated soil sites. The process is known as in situ vitrification (ISV). In situ vitrification is a remedial action process that destroys solid and liquid organic contaminants and incorporates radionuclides into a glass-like material that renders contaminants substantially less mobile and less likely to impact the environment. A large-scale operational acceptance test (LSOAT) was recently completed in which more than 180 t of vitrified soil were produced in each of three adjacent settings. The LSOAT demonstrated that the process conforms to the functional design criteria necessary for the large-scale radioactive test (LSRT) to be conducted following verification of the performance capabilities of the process. The energy requirements and vitrified block size, shape, and mass are sufficiently equivalent to those predicted by the ISV mathematical model to confirm its usefulness as a predictive tool. The LSOAT demonstrated an electrode replacement technique, which can be used if an electrode fails, and techniques have been identified to minimize air oxidation, thereby extending electrode life. A statistical analysis was employed during the LSOAT to identify graphite collars and an insulative surface as successful cold cap subsidence techniques. The LSOAT also showed that even under worst-case conditions, the off-gas system exceeds the flow requirements necessary to maintain a negative pressure on the hood covering the area being vitrified. The retention of simulated radionuclides and chemicals in the soil and off-gas system exceeds requirements so that projected emissions are one to two orders of magnitude below the maximum permissible concentrations of contaminants at the stack

  10. Audit of patient acceptance of nasal surgery as a day case procedure.

    Science.gov (United States)

    Tierney, P A; Samuel, D; Patel, K S; Thomas, D M

    1996-01-01

    A greater emphasis on day case surgery within the health service is seen as a method of improving efficiency and reducing expenditure. We interviewed 90 consecutive patients undergoing nasal surgery who had been preoperatively assessed as being fit for day case surgery. They were randomised into three groups regarding the duration of postoperative nasal packing. All patients stayed overnight following surgery and were interviewed prior to discharge. Some 52% of the overall sample would be happy to have nasal surgery performed as a day case. If the nasal pack was removed after two hours, this figure rose to 67%. This difference in patient acceptance did not attain statistical significance overall, but there was a significant difference in those undergoing submucosal resection. There was no difference in the age, sex distribution or type of surgery performed between each group. The audit commission quotes patient satisfaction with day case surgery at 80%. Nasal surgery was not examined in their report, but was included as one of a set of procedures suitable for consideration. Although day case nasal surgery may be safe, further research regarding patient acceptance is required.

  11. 10 CFR 36.41 - Construction monitoring and acceptance testing.

    Science.gov (United States)

    2010-01-01

    ... meets design specifications and generally accepted building code requirements for reinforced concrete... related alarm if used to meet § 36.59(b). For underwater irradiators, the licensee shall verify the...

  12. Laboratory procedures for waste form testing

    Energy Technology Data Exchange (ETDEWEB)

    Mast, E.S.

    1994-09-19

    The 100 and 300 areas of the Hanford Site are included on the US Environmental Protection Agencies (EPA) National Priorities List under the Comprehensive Environmental Response, Compensation, and Liability Act (CERCLA). Soil washing is a treatment process that is being considered for the remediation of the soil in these areas. Contaminated soil washing fines can be mixed or blended with cementations materials to produce stable waste forms that can be used for beneficial purposes in mixed or low-level waste landfills, burial trenches, environmental restoration sites, and other applications. This process has been termed co-disposal. The Co-Disposal Treatability Study Test Plan is designed to identify a range of cement-based formulations that could be used in disposal efforts in Hanford in co-disposal applications. The purpose of this document is to provide explicit procedural information for the testing of co-disposal formulations. This plan also provides a discussion of laboratory safety and quality assurance necessary to ensure safe, reproducible testing in the laboratory.

  13. 49 CFR 232.505 - Pre-revenue service acceptance testing plan.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 4 2010-10-01 2010-10-01 false Pre-revenue service acceptance testing plan. 232... § 232.505 Pre-revenue service acceptance testing plan. (a) General; submission of plan. Except as... its system the operating railroad or railroads shall submit a pre-revenue service acceptance...

  14. Acceptance test report, 241-SY-101 Flexible Receiver System, Phase 2 testing

    International Nuclear Information System (INIS)

    This document summarizes the results of the Phase 2 acceptance test of the 241-SY-101 Flexible Receiver System (FRS). The FRS is one of six major components of the Equipment Removal System, which has been designed to retrieve, transport, and store the test mixer pump currently installed in Tank 241-SY-101. The purpose of this acceptance test is to verify the strength of the containment bag and bag bottom cinching mechanism. It is postulated that 68 gallons of waste could be trapped inside the pump internals. The bag must be capable of supporting this waste if it shakes loose and drains to the bottom of the bag after the bag bottom has been cinched closed. This acceptance test was performed at the Maintenance and Storage Facility (MASF) Facility in the 400 area on January 23, 1995. The bag assembly supported the weight of 920 kg (2,020 lbs) of water with no leakage or damage to the bag. This value meets the acceptance criteria of 910 kg of water and therefore the results were found to be acceptable. The maximum volume of liquid expected to be held up in the pump internals is 258 L (68 gallons), which corresponds to 410 kg. This test weight gives just over a safety factor of 2. The bag also supported a small shock load while it was filled with water when the crane hoisted the bag assembly up and down. Based on the strength rating of the bag components, the bag assembly should support 2--3 times the test weight of 910 kg

  15. Utility-Scale Parabolic Trough Solar Systems: Performance Acceptance Test Guidelines, April 2009 - December 2010

    Energy Technology Data Exchange (ETDEWEB)

    Kearney, D.

    2011-05-01

    Prior to commercial operation, large solar systems in utility-size power plants need to pass a performance acceptance test conducted by the engineering, procurement, and construction (EPC) contractor or owners. In lieu of the present absence of ASME or other international test codes developed for this purpose, the National Renewable Energy Laboratory has undertaken the development of interim guidelines to provide recommendations for test procedures that can yield results of a high level of accuracy consistent with good engineering knowledge and practice. The Guidelines contained here are specifically written for parabolic trough collector systems with a heat-transport system using a high-temperature synthetic oil, but the basic principles are relevant to other CSP systems.

  16. The Careful Puppet Master: Reducing risk and fortifying acceptance testing with Jenkins CI

    Science.gov (United States)

    Smith, Jason A.; Richman, Gabriel; DeStefano, John; Pryor, James; Rao, Tejas; Strecker-Kellogg, William; Wong, Tony

    2015-12-01

    Centralized configuration management, including the use of automation tools such as Puppet, can greatly increase provisioning speed and efficiency when configuring new systems or making changes to existing systems, reduce duplication of work, and improve automated processes. However, centralized management also brings with it a level of inherent risk: a single change in just one file can quickly be pushed out to thousands of computers and, if that change is not properly and thoroughly tested and contains an error, could result in catastrophic damage to many services, potentially bringing an entire computer facility offline. Change management procedures can—and should—be formalized in order to prevent such accidents. However, like the configuration management process itself, if such procedures are not automated, they can be difficult to enforce strictly. Therefore, to reduce the risk of merging potentially harmful changes into our production Puppet environment, we have created an automated testing system, which includes the Jenkins CI tool, to manage our Puppet testing process. This system includes the proposed changes and runs Puppet on a pool of dozens of RedHat Enterprise Virtualization (RHEV) virtual machines (VMs) that replicate most of our important production services for the purpose of testing. This paper describes our automated test system and how it hooks into our production approval process for automatic acceptance testing. All pending changes that have been pushed to production must pass this validation process before they can be approved and merged into production.

  17. Flight test pilot evaluation of a delayed flap approach procedure

    Science.gov (United States)

    Bull, J. S.; Edwards, F. G.; Foster, J. D.; Hegarty, D. M.; Drinkwater, F. J., III

    1977-01-01

    Using NASA's CV-990 aircraft, a delayed flap approach procedure was demonstrated to nine guest pilots from the air transport industry. Four demonstration flights and 37 approaches were conducted under VFR weather conditions. A limited pilot evaluation of the delayed flap procedure was obtained from pilot comments and from questionaires they completed. Pilot acceptability, pilot workload, and ATC compatibility were quantitatively rated. The delayed flap procedure was shown to be feasible, and suggestions for further development work were obtained.

  18. 40 CFR 87.62 - Test procedure (propulsion engines).

    Science.gov (United States)

    2010-07-01

    ... PROGRAMS (CONTINUED) CONTROL OF AIR POLLUTION FROM AIRCRAFT AND AIRCRAFT ENGINES Test Procedures for Engine Exhaust Gaseous Emissions (Aircraft and Aircraft Gas Turbine Engines) § 87.62 Test procedure (propulsion... 40 Protection of Environment 20 2010-07-01 2010-07-01 false Test procedure (propulsion...

  19. 46 CFR 130.480 - Test procedure and operations manual.

    Science.gov (United States)

    2010-10-01

    ... 46 Shipping 4 2010-10-01 2010-10-01 false Test procedure and operations manual. 130.480 Section....480 Test procedure and operations manual. (a) A procedure for tests to be conducted on automated... detailed operations manual that describes the operation and indicates the location of each system...

  20. 40 CFR 401.13 - Test procedures for measurement.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 28 2010-07-01 2010-07-01 true Test procedures for measurement. 401.13... AND STANDARDS GENERAL PROVISIONS § 401.13 Test procedures for measurement. The test procedures for measurement which are prescribed at part 136 of this chapter shall apply to expressions of pollutant...

  1. Acceptance tests for electrometers used in radiotherapy; Pruebas de aceptacion para electrometros usados en radioterapia

    Energy Technology Data Exchange (ETDEWEB)

    Moreno-Torres, M.; Tornero-Lopez, A. M.; Ruiz-Arrebola, S.; Guirado, D.; Vilches, M.

    2015-05-01

    We present a set of tests designed to check the compliance of the requirements established in the standard IEC 60731:19972 and TRS 3981 technical report for reference class electrometers using the material normally available in a medical physics department. The characteristics under exam are: stability in the start-up, zero shift, zero drift, leakage current, linearity and repeatability. The method is applied to 4 electrometers: PTW UNIDOS, Sun Nuclear PC-Electrometer, Scanditronix-Wellhofer DOSE1 and Standard Imaging MAX4000. We checked that the electrometers comply with the requirements, although we detected some differences in the performance between them. The method proved itself precise enough and efficient to provide with reliable results for these tests and we consider it could be used to establish an internal procedure for the acceptance of this kind of equipment. (Author)

  2. Evaluation of a standard test method and material for low-activity waste product acceptance

    International Nuclear Information System (INIS)

    The dissolution behavior of a candidate reference glass has been measured under a range of test conditions. The data base from these tests can be used to evaluate the credibility and validity of test results reported by Private Contractors as part of the acceptance process for immobilized low-activity waste (ILAW) products for DOE wastes. The LRM-1 glass that was used in the tests was formulated to be compositionally representative of anticipated ILAW products for Hanford and other Department of Energy sites. Replicate tests with this glass were used to measure the variability in the response (i.e., the solution concentrations of B, Na, and Si) under different test conditions. The glass was further evaluated for possible use as a standard material by analysis of its composition, microstructure, density, and compressive strength. In addition, the Na leachability index was measured with the ANS 16.1 test, and the Toxicity Characteristic Leach Procedure was run. The results of those tests and analyses are summarized

  3. For your interest? The ethical acceptability of using non-invasive prenatal testing to test 'purely for information'.

    Science.gov (United States)

    Deans, Zuzana; Clarke, Angus J; Newson, Ainsley J

    2015-01-01

    Non-invasive prenatal testing (NIPT) is an emerging form of prenatal genetic testing that provides information about the genetic constitution of a foetus without the risk of pregnancy loss as a direct result of the test procedure. As with other prenatal tests, information from NIPT can help to make a decision about termination of pregnancy, plan contingencies for birth or prepare parents to raise a child with a genetic condition. NIPT can also be used by women and couples to test purely 'for information'. Here, no particular action is envisaged following the test; it is motivated entirely by an interest in the result. The fact that NIPT can be performed without posing a risk to the pregnancy could give rise to an increase in such requests. In this paper, we examine the ethical aspects of using NIPT 'purely for information', including the competing interests of the prospective parents and the future child, and the acceptability of testing for 'frivolous' reasons. Drawing on several clinical scenarios, we claim that arguments about testing children for genetic conditions are relevant to this debate. In addition, we raise ethical concerns over the potential for objectification of the child. We conclude that, in most cases, using NIPT to test for adult-onset conditions, carrier status or non-serious traits presenting in childhood would be unacceptable. PMID:25521970

  4. For your interest? The ethical acceptability of using non-invasive prenatal testing to test 'purely for information'.

    Science.gov (United States)

    Deans, Zuzana; Clarke, Angus J; Newson, Ainsley J

    2015-01-01

    Non-invasive prenatal testing (NIPT) is an emerging form of prenatal genetic testing that provides information about the genetic constitution of a foetus without the risk of pregnancy loss as a direct result of the test procedure. As with other prenatal tests, information from NIPT can help to make a decision about termination of pregnancy, plan contingencies for birth or prepare parents to raise a child with a genetic condition. NIPT can also be used by women and couples to test purely 'for information'. Here, no particular action is envisaged following the test; it is motivated entirely by an interest in the result. The fact that NIPT can be performed without posing a risk to the pregnancy could give rise to an increase in such requests. In this paper, we examine the ethical aspects of using NIPT 'purely for information', including the competing interests of the prospective parents and the future child, and the acceptability of testing for 'frivolous' reasons. Drawing on several clinical scenarios, we claim that arguments about testing children for genetic conditions are relevant to this debate. In addition, we raise ethical concerns over the potential for objectification of the child. We conclude that, in most cases, using NIPT to test for adult-onset conditions, carrier status or non-serious traits presenting in childhood would be unacceptable.

  5. SU-E-P-46: Clinical Acceptance Testing and Implementation of a Portable CT Unit

    Energy Technology Data Exchange (ETDEWEB)

    LaFrance, M; Marsh, S; Hicks, R; O’Donnell-Moran, G [Baystate Health Systems, Inc., Springfield, MA (United States)

    2015-06-15

    Purpose: Planning for the first installation in New England of a new portable CT unit to be used in the Operating Room required the integration of many departments including Surgery, Neurosurgery, Information Services, Clinical Engineering, Radiology and Medical Physics/Radiation Safety. Acceptance testing and the quality assurance procedures were designed to optimize image quality and patient and personnel radiation exposure. Methods: The vendor’s protocols were tested using the CT Dosimetry phantoms. The system displayed the CTDIw instead of the CTDIvol while testing the unit. Radiation exposure was compared to existing CT scanners from installed CT units throughout the facility. Brainlab measures all 4 periphery slots on the CT Dosimetry phantom. The ACR measures only the superior slot for the periphery measurement. A comprehensive radiation survey was also performed for several locations. Results: The CTDIvol measurements were comparable for the following studies: brain, C-Spine, and sinuses. However, the mobile CT measurements were slightly higher than other CT units but within acceptable tolerance if measured using the ACR method.Based on scatter measurements, it was determined if any personnel were to stay in the OR Suite during image acquisition that the appropriate lead apron and thyroid shields had to be worn.In addition, to reduce unnecessary scatter, there were two mobile 6 foot wide shields (1/16″ lead equivalent) available to protect personnel in the room and adjacent areas. Conclusion: Intraoperative CT provides the physician new opportunities for evaluation of the progression of surgical resections and device placement at the cost of increasing the amount of trained personnel required to perform this procedure. It also brings with it challenges to keep the radiation exposure to the patients and staff within reasonable limits.

  6. 42 CFR 426.500 - Procedure for filing an acceptable complaint concerning a provision (or provisions) of an NCD.

    Science.gov (United States)

    2010-10-01

    ... 42 Public Health 3 2010-10-01 2010-10-01 false Procedure for filing an acceptable complaint concerning a provision (or provisions) of an NCD. 426.500 Section 426.500 Public Health CENTERS FOR MEDICARE... Food and Drug Administration in relation to that drug or device. (d) Joint complaints—(1)...

  7. 42 CFR 426.400 - Procedure for filing an acceptable complaint concerning a provision (or provisions) of an LCD.

    Science.gov (United States)

    2010-10-01

    ... concerning a provision (or provisions) of an LCD. 426.400 Section 426.400 Public Health CENTERS FOR MEDICARE... COVERAGE DETERMINATIONS AND LOCAL COVERAGE DETERMINATIONS Review of an LCD § 426.400 Procedure for filing an acceptable complaint concerning a provision (or provisions) of an LCD. (a) The complaint....

  8. Medical versus surgical termination of early pregnancy: satisfaction with care, emotional impact and acceptability of the procedure

    Directory of Open Access Journals (Sweden)

    Prasanna L. Akkenapally

    2016-09-01

    Conclusions: Satisfaction with both the methods of medical and surgical abortion is high. Acceptability of the procedure next time was more with surgical abortion. MTOP had higher emotional impact. [Int J Reprod Contracept Obstet Gynecol 2016; 5(9.000: 3158-3166

  9. A review of acceptance testing of the Los Alamos, Canberra Alpha Sentry Continuous Air Monitor (CAM)

    International Nuclear Information System (INIS)

    Los Alamos National Laboratory (LANL) undertook the design and development of a new generation of alpha continuous air monitor (CAM) instrumentation that would incorporate advanced technologies in the design of the sampling inlet, multi-channel analyzer (MCA) electronics, solid state alpha detectors, radon background interference suppression, background interference compensation and based on spectral analysis, and microcomputer based data communication, processing, storage, and retrieval. The ANSI air monitoring instrument standards (Performance Specifications for Health Physics Instrumentation -- Occupational Airborne Radioactivity Monitoring Instrumentation, N42.17B) specify performance criteria and testing procedures for instruments and instrument systems designed to continuously sample and quantify airborne radioactivity in the workplace. Although the intent of the standard is to provide performance testing criteria for type testing, it is appropriate to evaluate the performance of a new instrument such as the Alpha Sentry against certain of these criteria for purposes of an acceptance test based on stated specifications and the Los Alamos CAM Requirements document. This report provides an overview of the results of these tests, as they pertain to instruments designed to detect alpha-emitting radionuclides in particulate form

  10. 49 CFR 238.111 - Pre-revenue service acceptance testing plan.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 4 2010-10-01 2010-10-01 false Pre-revenue service acceptance testing plan. 238... and General Requirements § 238.111 Pre-revenue service acceptance testing plan. (a) Passenger equipment that has previously been used in revenue service in the United States. For passenger...

  11. Equipment and procedures for triaxial testing of subgrade soils

    OpenAIRE

    Hormych, P.; Correia, A. Gomes; Gillett, S.

    1996-01-01

    One of the aims of the laboratory twinning project “A European Approach to Road Pavement Design” was to develop standard test procedures for repeated load triaxial testing of subgrade soils. For this purpose, repeatability tests were carried out to compare the triaxial equipment and test procedures of the four participating laboratories. This paper presents the results of these tests, and concludes by some recommendations for cyclic triaxial testing of subgrade soils.

  12. Test and acceptance from the AE point of view

    International Nuclear Information System (INIS)

    The power industry has undergone the transition from utilizing construction engineers for startup activities to utilizing test engineers who are responsible for the preparation or execution of a formal test program. This has come about because testing has been given sufficient importance to justify those participating within the industry to establish it as a specialty phase apart from engineering/design and construction. Presently testing is being conducted by those organizations that have either engineered and/or constructed the unit which may be a position of conflict in regards to unbiased test. The tester (third party) concept promotes the repetition, independence, and expertise of a test organization responsible to the utility for certifying that each system has been tested to the design criteria established by others. A move to this concept should result in better generating stations with higher availability because it has been completely tested by an organization with this sole contractual responsibility. The AE, NSSS, and utility would all benefit with possibly no additional costs incurred

  13. 40 CFR 60.123 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 6 2010-07-01 2010-07-01 false Test methods and procedures. 60.123 Section 60.123 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS... Smelters § 60.123 Test methods and procedures. (a) In conducting the performance tests required in §...

  14. Community acceptability of use of rapid diagnostic tests for malaria by community health workers in Uganda

    Directory of Open Access Journals (Sweden)

    Waiswa Peter

    2010-07-01

    Full Text Available Abstract Background Many malarious countries plan to introduce artemisinin combination therapy (ACT at community level using community health workers (CHWs for treatment of uncomplicated malaria. Use of ACT with reliance on presumptive diagnosis may lead to excessive use, increased costs and rise of drug resistance. Use of rapid diagnostic tests (RDTs could address these challenges but only if the communities will accept their use by CHWs. This study assessed community acceptability of the use of RDTs by Ugandan CHWs, locally referred to as community medicine distributors (CMDs. Methods The study was conducted in Iganga district using 10 focus group discussions (FGDs with CMDs and caregivers of children under five years, and 10 key informant interviews (KIIs with health workers and community leaders. Pre-designed FGD and KII guides were used to collect data. Manifest content analysis was used to explore issues of trust and confidence in CMDs, stigma associated with drawing blood from children, community willingness for CMDs to use RDTs, and challenges anticipated to be faced by the CMDs. Results CMDs are trusted by their communities because of their commitment to voluntary service, access, and the perceived effectiveness of anti-malarial drugs they provide. Some community members expressed fear that the blood collected could be used for HIV testing, the procedure could infect children with HIV, and the blood samples could be used for witchcraft. Education level of CMDs is important in their acceptability by the community, who welcome the use of RDTs given that the CMDs are trained and supported. Anticipated challenges for CMDs included transport for patient follow-up and picking supplies, adults demanding to be tested, and caregivers insisting their children be treated instead of being referred. Conclusion Use of RDTs by CMDs is likely to be acceptable by community members given that CMDs are properly trained, and receive regular technical

  15. CSI computer system/remote interface unit acceptance test results

    Science.gov (United States)

    Sparks, Dean W., Jr.

    1992-01-01

    The validation tests conducted on the Control/Structures Interaction (CSI) Computer System (CCS)/Remote Interface Unit (RIU) is discussed. The CCS/RIU consists of a commercially available, Langley Research Center (LaRC) programmed, space flight qualified computer and a flight data acquisition and filtering computer, developed at LaRC. The tests were performed in the Space Structures Research Laboratory (SSRL) and included open loop excitation, closed loop control, safing, RIU digital filtering, and RIU stand alone testing with the CSI Evolutionary Model (CEM) Phase-0 testbed. The test results indicated that the CCS/RIU system is comparable to ground based systems in performing real-time control-structure experiments.

  16. Shielding evaluation and acceptance testing of a prefabricated, modular, temporary radiation therapy treatment facility.

    Science.gov (United States)

    Ezzell, Gary A

    2004-01-01

    We have recently commissioned a temporary radiation therapy facility that is novel in two aspects: it was constructed using modular components, and the LINAC was installed in one of the modular sections before it was lifted into position. Additional steel and granular fill was added to the modular sections on-site during construction. The building will be disassembled and removed when no longer needed. This paper describes the radiation shielding specifications and survey of the facility, as well as the ramifications for acceptance testing occasioned by the novel installation procedure. The LINAC is a Varian 21EX operating at 6 MV and 18 MV. The radiation levels outside the vault satisfied the design criteria, and no anomalous leakage was detected along the joints of the modular structure. At 18 MV and 600 monitor units (MU) per minute, the radiation level outside the primary barrier walls was 8.5 micro Sv/h of photons; there were no detectable neutrons. Outside the direct-shielded door, the levels were 0.4 micro Sv/h of photons and 3.0 micro Sv/h of neutrons. The isocentricity of the accelerator met the acceptance criteria and was not affected by its preinstallation into an integrated baseframe and subsequent transport to the building site.

  17. Classroom Testing Procedures for University Students

    Institute of Scientific and Technical Information of China (English)

    WangLirong; YangFeng

    2004-01-01

    Testing and teaching are so closely related that it is virtually impossible to work in either area without being constantly concerned with the other. Testing must accompany English teaching and must reflect how much English know ledgestudents have grasped in a certain phase of English study. Tests should be constructed primarily to reinforce learning, to

  18. TMACS test procedure TP012: Panalarm software bridge

    International Nuclear Information System (INIS)

    This Test Procedure addresses the testing of the functionality of the Tank Monitor and Control System (TMACS) Panalarm bridge software. The features to be tested are: Bridge Initialization Options; Bridge Communication; Bridge Performance; Testing Checksum Errors; and Testing Command Reject Errors. Only the first three could be tested; the last two have been deferred to a later date

  19. A Procedure for Uncovering Acceptable and Nonacceptable Mortgage Applications through Discriminant Analysis Using Ranked Data

    OpenAIRE

    Larry G. Perry; Timothy P. Cronan; Donald A. Epley

    1987-01-01

    The procedure developed in this paper uses a less biased statistical technique than conventional discriminant analysis and parallels the ranking procedure used by loan officers. A variety of univariate and multivariate statistical procedures as well as comprehensive validation methods are used to develop a "best" model. The resulting model provides more accurate classification than other studies have shown, without violating federal law regarding discrimination.

  20. Acceptance Sampling Plans Based on Truncated Life Tests for Gompertz Distribution

    Directory of Open Access Journals (Sweden)

    Wenhao Gui

    2014-01-01

    Full Text Available An acceptance sampling plan for Gompertz distribution under a truncated life test is developed. For different acceptance numbers, consumer’s confidence levels and values of the ratio of the experimental time to the specified mean lifetime, the minimum sample sizes required to ensure the specified mean lifetime are obtained. The operating characteristic function values and the associated producer’s risks are also presented. An example is provided to illustrate the acceptance sampling plan.

  1. Some procedures for computerized ability testing

    NARCIS (Netherlands)

    Linden, van der Wim J.; Zwarts, Michel A.

    1989-01-01

    For computerized test systems to be operational, the use of item response theory is a prerequisite. As opposed to classical test theory, in item response models the abilities of the examinees and the properties of the items are parameterized separately. Hence, when measuring the abilities of examine

  2. 40 CFR 53.43 - Test procedures.

    Science.gov (United States)

    2010-07-01

    ... of isokinetic samplers. (ix) Calculate and record the coefficient of variation of the mass...) Calculate and record the coefficient of variation for the replicate sampling effectiveness measurements of... percent of the value specified in table D-2 and the variation at any test point in the test section...

  3. 16 CFR 1631.4 - Test procedure.

    Science.gov (United States)

    2010-01-01

    ...) Apparatus—(1) Test chamber. The test chamber shall consist of an open top hollow cube made of noncombustible... gel desiccant. (7) Gloves. Nonhygroscopic gloves (such as rubber or polyethylene) for handling the... Director of the Federal Register in accordance with 5 U.S.C. 552(a) and 1 CFR part 51. (2) Cutting....

  4. 16 CFR 1630.4 - Test procedure.

    Science.gov (United States)

    2010-01-01

    ...) Test chamber. The test chamber shall consist of an open top hollow cube made of noncombustible material...) Gloves. Nonhygroscopic gloves (such as rubber polyethylene) for handling the sample after drying, and... Director of the Federal Register in accordance with 5 U.S.C. 552(a) and 1 CFR part 51. (2) Cutting....

  5. Integrated Performance Testing Workshop - Supplemental Materials (Scripts and Procedures)

    Energy Technology Data Exchange (ETDEWEB)

    Baum, Gregory A.

    2014-02-01

    A variety of performance tests are described relating to: Material Transfers; Emergency Evacuation; Alarm Response Assessment; and an Enhanced Limited Scope Performance Test (ELSPT). Procedures are given for: nuclear material physical inventory and discrepancy; material transfers; and emergency evacuation.

  6. 40 CFR 87.5 - Special test procedures.

    Science.gov (United States)

    2010-07-01

    ...) CONTROL OF AIR POLLUTION FROM AIRCRAFT AND AIRCRAFT ENGINES General Provisions § 87.5 Special test... aircraft or aircraft engines, approve test procedures for any aircraft or aircraft engine that is...

  7. 21 CFR 1010.13 - Special test procedures.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 8 2010-04-01 2010-04-01 false Special test procedures. 1010.13 Section 1010.13 Food and Drugs FOOD AND DRUG ADMINISTRATION, DEPARTMENT OF HEALTH AND HUMAN SERVICES (CONTINUED) RADIOLOGICAL HEALTH PERFORMANCE STANDARDS FOR ELECTRONIC PRODUCTS: GENERAL Alternate Test Procedures §...

  8. Report on COTECH test procedure and characterization techniques

    DEFF Research Database (Denmark)

    Islam, Mohammad Aminul

    requirements for multifunctional integrated lighting rear lamp for cars (CRP) Section 7.Characterization techniques and test procedure requirements for micro-fresnel lenses for cell phone flash lights (HEPTAGON) Section 8.Characterization techniques and test procedure requirements for 5 Pin Ric Connector...

  9. 40 CFR 86.127-96 - Test procedures; overview.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Test procedures; overview. 86.127-96 Section 86.127-96 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS... determined by the procedure described in § 86.139. This testing requires a dilution tunnel as well as...

  10. HOT WATER COMFORT TEST PROCEDURE FOR SOLAR COMBISYSTEMS: PROPOSAL

    DEFF Research Database (Denmark)

    Furbo, Simon

    1999-01-01

    A proposal for a test procedure for hot water comfort for solar heating systems for combined space heating and domestic hot water supply was worked out.......A proposal for a test procedure for hot water comfort for solar heating systems for combined space heating and domestic hot water supply was worked out....

  11. 40 CFR 86.1840-01 - Special test procedures.

    Science.gov (United States)

    2010-07-01

    ... Compliance Provisions for Control of Air Pollution From New and In-Use Light-Duty Vehicles, Light-Duty Trucks... periodically regenerating trap oxidizer systems must propose a procedure for testing and certifying such... operation of the trap oxidizer system and the proposed certification and testing procedure. (d)...

  12. 40 CFR 600.009-85 - Hearing on acceptance of test data.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 29 2010-07-01 2010-07-01 false Hearing on acceptance of test data... Regulations for 1977 and Later Model Year Automobiles-General Provisions § 600.009-85 Hearing on acceptance of... orderly and expeditious manner. The parties may offer oral or written evidence, subject to the...

  13. Panel expertise for an Angoff standard setting procedure in progress testing : item writers compared to recently graduated students

    NARCIS (Netherlands)

    Verhoeven, BH; Verwijnen, GM; Muijtjens, AMM; Scherpbier, AJJA; van der Vleuten, CPM

    2002-01-01

    Introduction An earlier study showed that an Angoff procedure with greater than or equal to 10 recently graduated students as judges can be used to estimate the passing score of a progress test. As the acceptability and feasibility of this approach are questionable, we conducted an Angoff procedure

  14. Optimality of the Holm procedure among general step-down multiple testing procedures.

    Science.gov (United States)

    Gordon, Alexander Y; Salzman, Peter

    2008-09-15

    We study the class of general step-down multiple testing procedures, which contains the usually considered procedures determined by a nondecreasing sequence of thresholds (we call them threshold step-down, or TSD, procedures) as a parametric subclass. We show that all procedures in this class satisfying the natural condition of monotonicity and controlling the family-wise error rate (FWER) at a prescribed level are dominated by one of them - the classical Holm procedure. This generalizes an earlier result pertaining to the subclass of TSD procedures (Lehmann and Romano, Testing Statistical Hypotheses, 3rd ed., 2005). We also derive a relation between the levels at which a monotone step-down procedure controls the FWER and the generalized FWER (the probability of k or more false rejections).

  15. Affect and technology acceptance: A test of two mechanisms

    DEFF Research Database (Denmark)

    Scholderer, Joachim; Grunert, Klaus G.; Søndergaard, Helle Alsted

    are resistant to all forms of communicative interventions. A study is presented in which the effectiveness and mechanisms of direct-experience interventions are experimentally tested. Results indicate that the positive affect evoked by a single trial of a high-quality genetically modified food can lead......Commercialization of new technologies may be hampered by stakeholder resistance and a sceptical public. Genetic modification (GM) has suffered particularly from such problems. At present, for example, practically no products exist on the shelves of European retailers that are labelled as containing...... to strong changes in consumer attitudes (effect size d = .40). Furthermore, preferences for genetically modified products were found to be mainly dependent on the quality of the product participants had tried. Both effects were robust under contextual variations. Results are discussed in terms of theory...

  16. 40 CFR 204.54 - Test procedures.

    Science.gov (United States)

    2010-07-01

    ... to measure the sound level and to calculate the average sound level of portable air compressors on... discharge valve. The air discharge shall be piped clear of the test area or fed into an effective silencer. The sound pressure level due to the air discharge shall be at least 10 dB below the sound...

  17. 16 CFR 1616.5 - Test procedure.

    Science.gov (United States)

    2010-01-01

    ... holder over the center of the base of the cabinet at such a height that the bottom of the specimen is 1.7... Society for Testing and Materials, 1916 Race Street, Philadelphia, Pa. 19103). (10) Desiccator. An air... opening of the specimen holder. The bottom plate shall be the thickness of the specimen holder and the...

  18. A new approach to household appliance energy test procedures

    Energy Technology Data Exchange (ETDEWEB)

    Ernebrant, S.; Wihlborg, M.

    1999-03-01

    Energy test procedures provides the industry with a method to measure its products energy consumption. The energy test procedures are the technical foundation to every energy standard and labelling system. Depending on which country the product is going to be sold in, the manufacturers must follow different standards. This report concentrates on appliance test procedures, with the main focus on refrigerators. Recently, a new technology -microcontrollers- is undermining the credibility of the test procedures. New features, saving energy in real-life, are not picked up by the test procedures. It is estimated that as much as 30% of energy could be saved with this technology. Microcontrollers have also led to the possibility to circumvent tests. A new model is presented in the report, which handles these energy savings and make it harder cheating on tests. The model divides the test procedure in two parts; hardware tests and software tests, and uses a Matlab/Simulink computer model to calculate the energy consumption. Examples of hardware- and software test methods for refrigerators are described. A refrigerator is used as an example to present the model. The possibility to harmonize the energy standards to one global standard, which could mean substantial savings and make international trade more efficient, is also discussed 24 refs, 30 figs. Examination paper

  19. Test procedure for cation exchange chromatography

    International Nuclear Information System (INIS)

    The purpose of this test plan is to demonstrate the synthesis of inorganic antimonate ion exchangers and compare their performance against the standard organic cation exchangers. Of particular interest is the degradation rate of both inorganic and organic cation exchangers. This degradation rate will be tracked by determining the ion exchange capacity and thermal stability as a function of time, radiation dose, and chemical reaction

  20. Use of relative risk test to evaluate the influence of the brand on beer acceptability

    Directory of Open Access Journals (Sweden)

    Suzana Maria Della Lucia

    2014-02-01

    Full Text Available The acceptance of the product is very complex and it involves not only its sensory attributes but also non-sensory characteristics, often presented in the packaging or related to concepts regarding the consumer. The relative risk test was applied in order to investigate the effect of the brand, a non-sensory characteristic, on consumers’ acceptance of beer. Eight commercial brands of Pilsen beer were evaluated by 101 consumers in two sessions of acceptance tests: the blind test and the test with brand information. As a general view, it could be concluded that three brands of beer showed a positive impact over the sensory acceptance of the product and four brands negatively impacted on consumers’ evaluation. One of the brands did not show any influence over the consumers’ response. The relative risk method was practical and useful as it was easily calculated and reproduced, and its application permitted easy interpretation of the results.

  1. Quality assurance of radiopharmaceuticals-specifications and test procedures

    International Nuclear Information System (INIS)

    This report is a compilation of test methods used and specifications adopted for the Radiopharmaceutical Quality Assurance Test Programme conducted by the Australian Radiation Laboratory. In some cases test procedures described have been taken from various Pharmacopoeias or methods published in the literature. In other cases test methods have been developed at the ARL

  2. Results of the NASA Kennedy Space Center 50-MHz Doppler Radar Wind Profiler Operational Acceptance Test

    Science.gov (United States)

    Barbre', Robert E., Jr.; Decker, Ryan K.; Leahy, Frank B.; Huddleston, Lisa

    2016-01-01

    This paper presents results of the new Kennedy Space Center (KSC) 50-MHz Doppler Radar Wind Profiler (DRWP) Operational Acceptance Test (OAT). The goal of the OAT was to verify the data quality of the new DRWP against the performance of the previous DRWP in order to use wind data derived by the new DRWP for space launch vehicle operations support at the Eastern Range. The previous DRWP was used as a situational awareness asset for mission operations to identify rapid changes in the wind environment that weather balloons cannot depict. The Marshall Space Flight Center's Natural Environments Branch assessed data from the new DRWP collected during Jan-Feb 2015 against a specified set of test criteria. Data examination verified that the DRWP provides complete profiles every five minutes from 1.8-19.5 km in vertical increments of 150 m. Analysis of 49 concurrent DRWP and balloon profiles presented root mean square wind component differences around 2.0 m/s. Evaluation of the DRWP's coherence between five-minute wind pairs found the effective vertical resolution to be Nyquist-limited at 300 m for both wind components. In addition, the sensitivity to rejecting data that do not have adequate signal was quantified. This paper documents the data, quality control procedures, methodology, and results of each analysis.

  3. Test report for the run-in acceptance testing of the hydrogen mitigation retrieval Pump-3

    International Nuclear Information System (INIS)

    This report will provide the findings of the demonstration test conducted on the Double-Shell Tank (DST) 241-SY-101 HMR Pump-3 in accordance with WHC-SDWM-TP-434 ''Test plan for run-in acceptance testing of hydrogen mitigation/retrieval pump-3'' at the 400 Area Maintenance and Storage Facility (MASF) building from 7 June 1996 through 30 July 1996 per work package 4A-96-92/W. The DST 241-SY-101 hydrogen mitigation retrieval Pump-3 is a 200-HP submersible electric driven pump that has been modified for use in the DST 241-SY-101 containing mixed waste located in the 200W area. The pump has a motor driven rotation mechanism that allows the pump column to rotate through 355 degree. Prior to operation, pre-operational checks were performed which included loop calibration grooming and alignment of instruments, learning how plumb HMR-3 assembly hung in a vertical position and bump test of the motor to determine rotation direction. The pump was tested in the MASF Large Diameter Cleaning Vessel (LDCV) with process water at controlled temperatures and levels. In addition, the water temperature of the cooling water to the motor oil heat exchanger was recorded during testing. A 480-volt source powered a Variable Frequency Drive (VFD). The VFD powered the pump at various frequencies and voltages to control speed and power output of the pump. A second VFD powered the oil cooling pump. A third VFD was not available to operate the rotational drive motor during the 72 hour test, so it was demonstrated as operational before and after the test. A Mini Acquisition and Control System (Mini-DACS) controls pump functions and monitoring of the pump parameters. The Mini-DACS consists of three computers, software and some Programmable Logic Controllers (PLC). Startup and shutdown of either the pump motor or the oil cooling pump can be accomplished by the Mini-DACS. When the pump was in operation, the Mini-DACS monitors automatically collects data electronically. However, some required data

  4. Test report for run-in acceptance testing of hydrogen mitigation retrieval Pump-3

    Energy Technology Data Exchange (ETDEWEB)

    Berglin, B.G.

    1997-08-15

    This report will provide the findings of the demonstration test conducted on the Double-Shell Tank (DST) 241-SY-101 HMR Pump-3 in accordance with WHC-SDWM-TP-434 ``Test plan for run-in acceptance testing of hydrogen mitigation/retrieval pump-3`` at the 400 Area Maintenance and Storage Facility (MASF) building from 7 June 1996 through 30 July 1996 per work package 4A-96-92/W. The DST 241-SY-101 hydrogen mitigation retrieval Pump-3 is a 200-HP submersible electric driven pump that has been modified for use in the DST 241-SY-101 containing mixed waste located in the 200W area. The pump has a motor driven rotation mechanism that allows the pump column to rotate through 355{degree}. Prior to operation, pre-operational checks were performed which included loop calibration grooming and alignment of instruments, learning how plumb HMR-3 assembly hung in a vertical position and bump test of the motor to determine rotation direction. The pump was tested in the MASF Large Diameter Cleaning Vessel (LDCV) with process water at controlled temperatures and levels. In addition, the water temperature of the cooling water to the motor oil heat exchanger was recorded during testing. A 480-volt source powered a Variable Frequency Drive (VFD). The VFD powered the pump at various frequencies and voltages to control speed and power output of the pump. A second VFD powered the oil cooling pump. A third VFD was not available to operate the rotational drive motor during the 72 hour test, so it was demonstrated as operational before and after the test. A Mini Acquisition and Control System (Mini-DACS) controls pump functions and monitoring of the pump parameters. The Mini-DACS consists of three computers, software and some Programmable Logic Controllers (PLC). Startup and shutdown of either the pump motor or the oil cooling pump can be accomplished by the Mini-DACS. When the pump was in operation, the Mini-DACS monitors automatically collects data electronically. However, some required data

  5. [Voluntary testing procedures of farm animal housing equipment according to the Animal Welfare Act of 1998].

    Science.gov (United States)

    Hesse, D; Knierim, U; von Borell, E; Herrmann, H; Koch, L; Müller, C; Rauch, H W; Sachser, N; Schwabenbauer, K; Zerbe, F

    1999-04-01

    Before its broad application in practice, housing equipment should be tested, in particular with regard to animal welfare. The differing positions of the German Federal Council (Bundesrat) and the German Federal Parliament (Bundestag), whether such testing should be mandatory or voluntary, have been conciliated in the amended animal welfare act by empowering the Federal Ministry of Food, Agriculture and Fisheries (BML) to fix official standards for voluntary testing procedures by regulation. On request of the BML, a report as scientific basis for a draft regulation is currently prepared by the scientific animal welfare committee of the German Agricultural Society (DLG). The scientific animal welfare committee has been appointed by the DLG in order to provide support in the effort to strengthen animal welfare aspects in the DLG-utility testing procedure of housing equipment, which is in place since 1953. The committee elaborates standards concerning testing methods, assessment criteria and the necessary size of investigations. As required, the scientific animal welfare committee may support the DLG-testing bodies in the implementation of the animal welfare part of the testing procedure. It will, moreover, be involved in the welfare assessment based on the testing results. The amendments of the already established testing procedure will help to fulfill the general requirements on an acceptable animal welfare testing procedure. While keeping in mind that there are certain limits in what can be achieved by a voluntary testing procedure, the enhanced consideration of animal welfare aspects within the DLG-utility testing procedure has the advantage to be relatively unbureaucratic and in line with EU legislation, and is, therefore, an appropriate tool for a contibrution to improved animal welfare in livestock housing.

  6. Full-Scale Cask Testing and Public Acceptance of Spent Nuclear Fuel Shipments - 12254

    Energy Technology Data Exchange (ETDEWEB)

    Dilger, Fred [Black Mountain Research, Henderson, NV 81012 (United States); Halstead, Robert J. [State of Nevada Agency for Nuclear Projects Carson City, NV 80906 (United States); Ballard, James D. [Department of Sociology, California State University, Northridge Northridge, CA 91330 (United States)

    2012-07-01

    Full-scale physical testing of spent fuel shipping casks has been proposed by the National Academy of Sciences (NAS) 2006 report on spent nuclear fuel transportation, and by the Presidential Blue Ribbon Commission (BRC) on America's Nuclear Future 2011 draft report. The U.S. Nuclear Regulatory Commission (NRC) in 2005 proposed full-scale testing of a rail cask, and considered 'regulatory limits' testing of both rail and truck casks (SRM SECY-05-0051). The recent U.S. Department of Energy (DOE) cancellation of the Yucca Mountain project, NRC evaluation of extended spent fuel storage (possibly beyond 60-120 years) before transportation, nuclear industry adoption of very large dual-purpose canisters for spent fuel storage and transport, and the deliberations of the BRC, will fundamentally change assumptions about the future spent fuel transportation system, and reopen the debate over shipping cask performance in severe accidents and acts of sabotage. This paper examines possible approaches to full-scale testing for enhancing public confidence in risk analyses, perception of risk, and acceptance of spent fuel shipments. The paper reviews the literature on public perception of spent nuclear fuel and nuclear waste transportation risks. We review and summarize opinion surveys sponsored by the State of Nevada over the past two decades, which show consistent patterns of concern among Nevada residents about health and safety impacts, and socioeconomic impacts such as reduced property values along likely transportation routes. We also review and summarize the large body of public opinion survey research on transportation concerns at regional and national levels. The paper reviews three past cask testing programs, the way in which these cask testing program results were portrayed in films and videos, and examines public and official responses to these three programs: the 1970's impact and fire testing of spent fuel truck casks at Sandia National

  7. Full-Scale Cask Testing and Public Acceptance of Spent Nuclear Fuel Shipments - 12254

    International Nuclear Information System (INIS)

    Full-scale physical testing of spent fuel shipping casks has been proposed by the National Academy of Sciences (NAS) 2006 report on spent nuclear fuel transportation, and by the Presidential Blue Ribbon Commission (BRC) on America's Nuclear Future 2011 draft report. The U.S. Nuclear Regulatory Commission (NRC) in 2005 proposed full-scale testing of a rail cask, and considered 'regulatory limits' testing of both rail and truck casks (SRM SECY-05-0051). The recent U.S. Department of Energy (DOE) cancellation of the Yucca Mountain project, NRC evaluation of extended spent fuel storage (possibly beyond 60-120 years) before transportation, nuclear industry adoption of very large dual-purpose canisters for spent fuel storage and transport, and the deliberations of the BRC, will fundamentally change assumptions about the future spent fuel transportation system, and reopen the debate over shipping cask performance in severe accidents and acts of sabotage. This paper examines possible approaches to full-scale testing for enhancing public confidence in risk analyses, perception of risk, and acceptance of spent fuel shipments. The paper reviews the literature on public perception of spent nuclear fuel and nuclear waste transportation risks. We review and summarize opinion surveys sponsored by the State of Nevada over the past two decades, which show consistent patterns of concern among Nevada residents about health and safety impacts, and socioeconomic impacts such as reduced property values along likely transportation routes. We also review and summarize the large body of public opinion survey research on transportation concerns at regional and national levels. The paper reviews three past cask testing programs, the way in which these cask testing program results were portrayed in films and videos, and examines public and official responses to these three programs: the 1970's impact and fire testing of spent fuel truck casks at Sandia National Laboratories, the 1980's

  8. Biaxial testing for fabrics and foils optimizing devices and procedures

    CERN Document Server

    Beccarelli, Paolo

    2015-01-01

    This book offers a well-structured, critical review of current design practice for tensioned membrane structures, including a detailed analysis of the experimental data required and critical issues relating to the lack of a set of design codes and testing procedures. The technical requirements for biaxial testing equipment are analyzed in detail, and aspects that need to be considered when developing biaxial testing procedures are emphasized. The analysis is supported by the results of a round-robin exercise comparing biaxial testing machines that involved four of the main research laboratories in the field. The biaxial testing devices and procedures presently used in Europe are extensively discussed, and information is provided on the design and implementation of a biaxial testing rig for architectural fabrics at Politecnico di Milano, which represents a benchmark in the field. The significance of the most recent developments in biaxial testing is also explored.

  9. 40 CFR 600.111-08 - Test procedures.

    Science.gov (United States)

    2010-07-01

    ... ECONOMY AND CARBON-RELATED EXHAUST EMISSIONS OF MOTOR VEHICLES Fuel Economy Regulations for 1978 and Later...) Highway fuel economy testing procedures. (1) The Highway Fuel Economy Dynamometer Procedure (HFET..., carbon dioxide using a constant volume (variable dilution) sampler. Diesel dilute exhaust is...

  10. Acceptability of HIV counselling and testing among tuberculosis patients in south Ethiopia

    Directory of Open Access Journals (Sweden)

    Endale Aschalew

    2007-05-01

    Full Text Available Abstract Background To benefit from available care and treatment options, patients should first be counselled and tested for HIV. Our aim was to assess the acceptability of HIV testing among tuberculosis patients under routine care conditions in south Ethiopia. Methods We interviewed all adult tuberculosis patients who were treated at Arba Minch Hospital in Ethiopia between January and August 2005. After recording socio-demographic information and tuberculosis treatment history, we referred those patients who showed initial willingness to a counsellor for HIV counselling and testing. Rapid test methods were used following a pretest counselling session. The results were disclosed during a post-test counselling session. We used the logistic regression method to assess factors associated with willingness and acceptability. Results 190 adult tuberculosis patients were treated at the hospital and all of them consented to take part in the study. Their median age was 30 years (range, 15–68 and 52% of them were males. 49 patients (26% were previously tested including 29 (59% HIV positive. Of 161 patients (excluding the 29 already positive, 118 (73% were willing to be tested and 58% (68/118 of those willing accepted the test. The overall acceptability rate was 35% (56/161. Fourteen (20.6% were HIV positive and women were more likely to be HIV infected (p = 0.029. Unemployment and self-perceived high risk of HIV infection were associated with initial willingness (OR [95%CI]:2.6 [1.3–5.5] vs. 5.0 [1.1–22.4], respectively. However, only being unemployed was associated with accepting the test (OR = 4.2; 95%CI = 1.9–9.3. Conclusion The low acceptability of HIV counselling and testing among tuberculosis patients poses a challenge to the scale-up of TB/HIV collaborative efforts. There is a need for alternative counselling and testing strategies.

  11. Test Procedure - pumping system for caustic addition project

    International Nuclear Information System (INIS)

    This test procedure provides the requirements for sub-system testing and integrated operational testing of the submersible mixer pump and caustic addition equipment by WHC and Kaiser personnel at the Rotating Equipment Shop run-in pit (Bldg. 272E)

  12. 49 CFR 572.163 - Neck assembly and test procedure.

    Science.gov (United States)

    2010-10-01

    ... assembly is assembled and tested as specified in 49 CFR 572.123 (Subpart N). ... 49 Transportation 7 2010-10-01 2010-10-01 false Neck assembly and test procedure. 572.163 Section... TRAFFIC SAFETY ADMINISTRATION, DEPARTMENT OF TRANSPORTATION (CONTINUED) ANTHROPOMORPHIC TEST...

  13. Specifications for Testing Procedures at Svåheia

    DEFF Research Database (Denmark)

    Margheritini, Lucia; Kofoed, Jens Peter

    This report is realized in order to define testing procedures of the SSG pilot in Svåheia location. This will be done by listing all the relevant parameters related to the structure performance.......This report is realized in order to define testing procedures of the SSG pilot in Svåheia location. This will be done by listing all the relevant parameters related to the structure performance....

  14. CEL Working procedures for WRAP 2A formulation development test

    International Nuclear Information System (INIS)

    The WRAP 2A facility will encapsulate retrieved, stored, and newly generated contact-handled mixed low level waste (MLLW) into 55-500 gal cementitous forms. Standardized test procedures will be required to facilitate this process. Cementitous specimens will be prepared from simulated drum wastes and will be tested in the Chemical Engineering Laboratory using the laboratory operating/working procedures encorporated into this document

  15. Lagoon Seepage Testing Procedures for Central Facilities Area (CFA) Sewage Lagoons at Idaho National Laboratory Butte County, Idaho April 2014

    Energy Technology Data Exchange (ETDEWEB)

    Alan Giesbrecht

    2014-05-01

    The lagoon seepage testing procedures are documented herein as required by the Wastewater Rules (IDAPA 58.01.16.493). The Wastewater Rules and Wastewater Reuse Permit LA-000141-03 require that the procedure used for performing a seepage test be approved by IDEQ prior to conducting the seepage test. The procedures described herein are based on a seepage testing plan that was developed by J-U-B ENGINEERS, Inc. (J-U-B) and has been accepted by several IDEQ offices for lagoons in Idaho.

  16. Understanding Student Teachers’ Behavioural Intention to Use Technology: Technology Acceptance Model (TAM) Validation and Testing

    OpenAIRE

    Kung-Teck, Wong; Rosma bt Osman; Pauline Swee Choo, Goh; Mohd Khairezan Rahmat

    2013-01-01

    This study sets out to validate and test the Technology Acceptance Model (TAM) in the context of Malaysian student teachers’ integration of their technology in teaching and learning. To establish factorial validity, data collected from 302 respondents were tested against the TAM using confirmatory factor analysis (CFA), and structural equation modelling (SEM) was used for model comparison and hypotheses testing. The goodness-of-fit test of the analysis shows partial support of the applicabili...

  17. Understanding Student Teachers' Behavioural Intention to Use Technology: Technology Acceptance Model (TAM) Validation and Testing

    Science.gov (United States)

    Wong, Kung-Teck; Osman, Rosma bt; Goh, Pauline Swee Choo; Rahmat, Mohd Khairezan

    2013-01-01

    This study sets out to validate and test the Technology Acceptance Model (TAM) in the context of Malaysian student teachers' integration of their technology in teaching and learning. To establish factorial validity, data collected from 302 respondents were tested against the TAM using confirmatory factor analysis (CFA), and structural equation…

  18. 40 CFR 1065.520 - Pre-test verification procedures and pre-test data collection.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 32 2010-07-01 2010-07-01 false Pre-test verification procedures and pre-test data collection. 1065.520 Section 1065.520 Protection of Environment ENVIRONMENTAL PROTECTION... Specified Duty Cycles § 1065.520 Pre-test verification procedures and pre-test data collection. (a) If...

  19. Round-robin testing of Soleq EV cort according to the SAE J1634 test procedure dated May 1993

    Energy Technology Data Exchange (ETDEWEB)

    Cole, G.H.

    1993-05-01

    The Society of Automotive Engineers Recommended Practice, SAE J1634, {open_quotes}Electric Vehicle Energy Consumption and Range Test Procedure{close_quotes}, May 1993, describes a standard method of determining the range and energy consumption for electric vehicles. Consequent to the U.S. Department of Energy`s (DOE) rulemaking released on February 4, 1994, the EPA is currently considering factoring electric vehicles into the Corporate Average Fuel Economy (CAFE) calculations using the SAE J1634 procedure. The purpose of this project is to provide information regarding the suitability of this recommended practice for determining the energy economy value to be factored into the CAFE. Issues leading to possible inconsistent results (such as the repeatability of the procedure, thoroughness of the procedure`s methods, and variance between different laboratories using different dynamometers) need to be resolved prior to passing legislation which will mandate use of this test in determining the electric vehicle CAFE credit. To this end, separate tests were performed on a Soleq EVcort vehicle by the INEL, the EPA, Ford Motor Company, Southwest Research Institute, and the California Air Resources Board using their own facilities and personnel. Acceptable departures from the driving profile prescribed by SAEJ1634 are not well defined. This deficiency in the procedure is even more noticeable due to the EVcort`s marginal acceleration performance.

  20. European wind turbine procedure development blade test methods and techniques

    Energy Technology Data Exchange (ETDEWEB)

    Bulder, B.H.; Dam, J.J.D. van; Delft, D.R.V. van [and others

    1999-03-01

    In this paper the preliminary results obtained by performing the second task of the `European Wind Turbine Testing Procedure Development` project are presented. This project is performed within and with financial support of the Standards, Measurements and Testing programme of the European Commission. (au)

  1. 40 CFR 86.608-98 - Test procedures.

    Science.gov (United States)

    2010-07-01

    ... part for any motor vehicle which is not susceptible to satisfactory testing using the procedures in... testing diesel vehicles, or methanol-fueled Otto-cycle vehicles, the manufacturer shall allow a minimum of.... When not in use, the chopper motors of the infrared analyzers are turned off and the phototube...

  2. 16 CFR 1512.18 - Tests and test procedures.

    Science.gov (United States)

    2010-01-01

    ... 1512.18 Commercial Practices CONSUMER PRODUCT SAFETY COMMISSION FEDERAL HAZARDOUS SUBSTANCES ACT... load using the least square method of obtaining the best straight line curve. (3) Performance test. The... shall be open at least 360° from a finger tight condition. A separation force of at least 111 N (25...

  3. High acceptance of an early dyslexia screening test involving genetic analyses in Germany.

    Science.gov (United States)

    Wilcke, Arndt; Müller, Bent; Schaadt, Gesa; Kirsten, Holger; Boltze, Johannes

    2016-02-01

    Dyslexia is a developmental disorder characterized by severe problems in the acquisition of reading and writing skills. It has a strong neurobiological basis. Genetic influence is estimated at 50-70%. One of the central problems with dyslexia is its late diagnosis, normally not before the end of the 2nd grade, resulting in the loss of several years for early therapy. Currently, research is focusing on the development of early tests for dyslexia, which may be based on EEG and genetics. Our aim was to determine the acceptance of such a future test among parents. We conducted a representative survey in Germany with 1000 parents of children aged 3-7 years, with and without experience of dyslexia. 88.7% of the parents supported the introduction of an early test for dyslexia based on EEG and genetics; 82.8% would have their own children tested, and 57.9% were willing to pay for the test if health insurance did not cover the costs. Test acceptance was significantly higher if parents had prior experience with dyslexia. The perceived benefits of such a test were early recognition and remediation and, preventing deficits. Concerns regarded the precision of the test, its potentially stigmatizing effect and its costs. The high overall support for the test leads to the conclusion that parents would accept a test for dyslexia based on EEG and genetics. PMID:26036858

  4. High acceptance of an early dyslexia screening test involving genetic analyses in Germany.

    Science.gov (United States)

    Wilcke, Arndt; Müller, Bent; Schaadt, Gesa; Kirsten, Holger; Boltze, Johannes

    2016-02-01

    Dyslexia is a developmental disorder characterized by severe problems in the acquisition of reading and writing skills. It has a strong neurobiological basis. Genetic influence is estimated at 50-70%. One of the central problems with dyslexia is its late diagnosis, normally not before the end of the 2nd grade, resulting in the loss of several years for early therapy. Currently, research is focusing on the development of early tests for dyslexia, which may be based on EEG and genetics. Our aim was to determine the acceptance of such a future test among parents. We conducted a representative survey in Germany with 1000 parents of children aged 3-7 years, with and without experience of dyslexia. 88.7% of the parents supported the introduction of an early test for dyslexia based on EEG and genetics; 82.8% would have their own children tested, and 57.9% were willing to pay for the test if health insurance did not cover the costs. Test acceptance was significantly higher if parents had prior experience with dyslexia. The perceived benefits of such a test were early recognition and remediation and, preventing deficits. Concerns regarded the precision of the test, its potentially stigmatizing effect and its costs. The high overall support for the test leads to the conclusion that parents would accept a test for dyslexia based on EEG and genetics.

  5. Local tolerance testing under REACH: Accepted non-animal methods are not on equal footing with animal tests.

    Science.gov (United States)

    Sauer, Ursula G; Hill, Erin H; Curren, Rodger D; Raabe, Hans A; Kolle, Susanne N; Teubner, Wera; Mehling, Annette; Landsiedel, Robert

    2016-07-01

    In general, no single non-animal method can cover the complexity of any given animal test. Therefore, fixed sets of in vitro (and in chemico) methods have been combined into testing strategies for skin and eye irritation and skin sensitisation testing, with pre-defined prediction models for substance classification. Many of these methods have been adopted as OECD test guidelines. Various testing strategies have been successfully validated in extensive in-house and inter-laboratory studies, but they have not yet received formal acceptance for substance classification. Therefore, under the European REACH Regulation, data from testing strategies can, in general, only be used in so-called weight-of-evidence approaches. While animal testing data generated under the specific REACH information requirements are per se sufficient, the sufficiency of weight-of-evidence approaches can be questioned under the REACH system, and further animal testing can be required. This constitutes an imbalance between the regulatory acceptance of data from approved non-animal methods and animal tests that is not justified on scientific grounds. To ensure that testing strategies for local tolerance testing truly serve to replace animal testing for the REACH registration 2018 deadline (when the majority of existing chemicals have to be registered), clarity on their regulatory acceptance as complete replacements is urgently required.

  6. Quality Assurance Plan for the AL3 Test Procedure

    CERN Document Server

    Béjar-Alonso, Isabel

    1999-01-01

    This paper describes the new quality assurance plan for the Alarms-of-Level-3 (AL3) test. The aim of the plan is to introduce engineering techniques and to standardise and simplify the procedures for carrying out tests following Safety Instruction 37 (IS37). The procedures are to co-ordinate all the services involved (fire brigade, maintenance and computer support) and to create a consistent documentation. When the procedures are implemented, it will be possible to determine with confidence how field actions are carried out and to measure actual performance. The focus will be on personnel training and documentation. It is important however to keep documentation and procedures to a reasonable level that can be maintained at appropriate intervals. The plan is the result of an internal requirement from ST/MC and a formal request from Installations Nucléaires de Base (INB).

  7. Nevada Test Site Waste Acceptance Criteria (NTSWAC), Rev. 7-01

    Energy Technology Data Exchange (ETDEWEB)

    NSTec Environmental Management

    2009-05-01

    This document establishes the U.S. Department of Energy, National Nuclear Security Administration Nevada Site Office, Nevada Test Site Waste Acceptance Criteria (NTSWAC). The NTSWAC provides the requirements, terms, and conditions under which the Nevada Test Site (NTS) will accept low-level radioactive waste and mixed low-level waste for disposal. The NTSWAC includes requirements for the generator waste certification program, characterization, traceability, waste form, packaging, and transfer. The criteria apply to radioactive waste received at the NTS Area 3 and Area 5 Radioactive Waste Management Complex for disposal.

  8. W-026 acceptance test plan plant control system hardware (submittal {number_sign} 216)

    Energy Technology Data Exchange (ETDEWEB)

    Watson, T.L., Fluor Daniel Hanford

    1997-02-14

    Acceptance Testing of the WRAP 1 Plant Control System Hardware will be conducted throughout the construction of WRAP I with the final testing on the Process Area hardware being completed in November 1996. The hardware tests will be broken out by the following functional areas; Local Control Units, Operator Control Stations in the WRAP Control Room, DMS Server, PCS Server, Operator Interface Units, printers, DNS terminals, WRAP Local Area Network/Communications, and bar code equipment. This document will contain completed copies of each of the hardware tests along with the applicable test logs and completed test exception reports.

  9. Testing Procedure for the Single Fiber Fragmentation Test

    DEFF Research Database (Denmark)

    Feih, Stefanie; Wonsyld, Karen; Minzari, Daniel;

    This report describes the details of the single fiber fragmentation test as conducted at the materials research department (AFM) at Risø. The equipment and specimen manufacture is described in detail. Furthermore, examples of results interpretation are given. For the experiments in this report......, specimens with one E-glass fiber placed inside an epoxy or polyester matrix were used. Elongating the specimens with a mini tensile tester, which was placed under a microscope, leads to fiber fragmentations. Different bonding strengths between fiber and matrix result in differences in the critical fracture...

  10. Accepting test-anxiety-related thoughts increases academic performance among undergraduate students

    Directory of Open Access Journals (Sweden)

    Senay Ibrahim

    2012-01-01

    Full Text Available Given that unwanted thoughts are enhanced when suppressed, we tested among college freshmen who were about to take an academic exam if an acceptance strategy consisting of not suppressing intrusive thoughts will improve test performance. This strategy proved superior to students’ own default strategies as much as a modified, alternative strategy, avoiding the antecedents of intrusive thoughts. Moreover, the combination of the two strategies counteracted a stronger, negative effect of test anxiety on test performance as compared with each strategy used alone. The results suggest that not only intrusive thoughts per se but also the suppression of these thoughts can disrupt test performance, and hint that approaching such thoughts with acceptance may not interfere with simultaneously working toward avoiding the antecedents of these thoughts.

  11. Utility-Scale Power Tower Solar Systems: Performance Acceptance Test Guidelines

    Energy Technology Data Exchange (ETDEWEB)

    Kearney, D.

    2013-03-01

    The purpose of these Guidelines is to provide direction for conducting performance acceptance testing for large power tower solar systems that can yield results of a high level of accuracy consistent with good engineering knowledge and practice. The recommendations have been developed under a National Renewable Energy Laboratory (NREL) subcontract and reviewed by stakeholders representing concerned organizations and interests throughout the concentrating solar power (CSP) community. An earlier NREL report provided similar guidelines for parabolic trough systems. These Guidelines recommend certain methods, instrumentation, equipment operating requirements, and calculation methods. When tests are run in accordance with these Guidelines, we expect that the test results will yield a valid indication of the actual performance of the tested equipment. But these are only recommendations--to be carefully considered by the contractual parties involved in the Acceptance Tests--and we expect that modifications may be required to fit the particular characteristics of a specific project.

  12. European wind turbine testing procedure developments. Task 2: Power quality

    DEFF Research Database (Denmark)

    Sørensen, Poul Ejnar; Friis Pedersen, Troels; Gerdes, G.;

    2001-01-01

    The present report describes the work done in the power quality subtask of the European Wind Turbine Testing Procedure Developments project funded by the EU SMT program. The objective of the power quality subtask has been to make recommendations andprovide background for new standards...... for measurement and testing of wind turbine power quality. The focus in the work has been to support the ongoing standardisation work in IEC with a new standard IEC61400-21 for measurement and assessment of powerquality characteristics of grid connected wind turbines. The work has also been based on the power...... quality measuremnet procedure in the Measnet cooperation of European test stations for wind turbines. The first working item of the project has been toverify the state of the art of the measurement procedures by analyses and comparisons of the measurements and data processing software of the participating...

  13. Operability test procedure [Tank] 241-SY-101 equipment removal system

    International Nuclear Information System (INIS)

    The 241-SY-101 equipment removal system (ERS) consists of components, equipment, instrumentation and procedures that will provide the means to disconnect, retrieve, contain, load and transport the Mitigation Pump Assembly (MPA) from waste Tank 241-SY-101 to the Central Waste Complex (CWC). The Operability Test Procedure (OTP) will test the interfaces between ERS components and will rehearse the procedure for MPA removal and transportation to the extent they can be mocked-up at the CTF (Cold Test Facility). At the conclusion of the OTP, the ERS components and equipment will be removed from the CTF, entered into the Component Based Recall System (CBRS), and stored until needed for actual MPA removal and transportation

  14. A new general pressure-analysis procedure for slug tests

    Energy Technology Data Exchange (ETDEWEB)

    Peres, A.M.M.; Onur, M.; Reynolds, A.C. (Univ. of Tulsa, OK (United States))

    1993-12-01

    A new analysis procedure for determining formation flow capacity and skin factor from slug-test data is presented. The procedure arises from exact deconvolution equations that convert measured slug-test pressure data into equivalent pressure and pressure-derivative responses that would be obtained if the well were produced at a constant surface flow rate. The converted data then can be analyzed by use of existing wellbore-storage and skin type curves for the particular reservoir or well model represented by the field data. For cases where the slug test is short, the authors show that flow rate convolution can be incorporated to improve analysis reliability. The analysis procedures do not require direct knowledge of the sandface flow rate.

  15. Acceptance Test Report for the Modular Automation System (MAS) Manufactured by Honeywell Inc

    International Nuclear Information System (INIS)

    This document details the performance of the acceptance test of the Honeywell MAS Control System for equipment to be installed in gloveboxes HA-20MB and HA-211 at a later date. Equipment that was anticipated included 6 stabilization furnaces, only three and their associated equipment were installed

  16. Integrating Telemedicine for Disaster Response: Testing the Emergency Telemedicine Technology Acceptance Model

    Science.gov (United States)

    Davis, Theresa M.

    2013-01-01

    Background: There is little evidence that technology acceptance is well understood in healthcare. The hospital environment is complex and dynamic creating a challenge when new technology is introduced because it impacts current processes and workflows which can significantly affect patient care delivery and outcomes. This study tested the effect…

  17. 40 CFR 86.1830-01 - Acceptance of vehicles for emission testing.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 19 2010-07-01 2010-07-01 false Acceptance of vehicles for emission testing. 86.1830-01 Section 86.1830-01 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY... vehicles shall either be randomly selected production parts or parts verified to be in the middle...

  18. A FORTRAN IV Program for Multiple-choice Tests with Predetermined Minimal Acceptable Performance Levels

    Science.gov (United States)

    Noe, Michael J.

    1976-01-01

    A Fortran IV multiple choice test scoring program for an IBM 370 computer is described that computes minimally acceptable performance levels and compares student scores to these levels. The program accomodates up to 500 items with no more than nine alternatives from a group of examinees numbering less than 10,000. (Author)

  19. Accepting test-anxiety-related thoughts increases academic performance among undergraduate students

    OpenAIRE

    Senay Ibrahim; Cetinkaya Mustafa; Usak Muhammet

    2012-01-01

    Given that unwanted thoughts are enhanced when suppressed, we tested among college freshmen who were about to take an academic exam if an acceptance strategy consisting of not suppressing intrusive thoughts will improve test performance. This strategy proved superior to students’ own default strategies as much as a modified, alternative strategy, avoiding the antecedents of intrusive thoughts. Moreover, the combination of the two strategies counteracted a stronger, negative effect of te...

  20. Empirical testing of forecast update procedure forseasonal products

    DEFF Research Database (Denmark)

    Wong, Chee Yew; Johansen, John

    2008-01-01

    a toy supply chain. The theoretical simulation involves historical weekly consumer demand data for 122 toy products. The empirical test is then carried out in real-time with 291 toy products. The results show that the proposed forecast updating procedure: 1) reduced forecast errors of the annual...

  1. 40 CFR 60.534 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 6 2010-07-01 2010-07-01 false Test methods and procedures. 60.534 Section 60.534 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS... concentrations may be measured with either: (1) Method 5G, if a dilution tunnel sampling location is used, or...

  2. 40 CFR 60.166 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 6 2010-07-01 2010-07-01 false Test methods and procedures. 60.166 Section 60.166 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS (CONTINUED) STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES Standards of Performance for Primary...

  3. 16 CFR 1700.20 - Testing procedure for special packaging.

    Science.gov (United States)

    2010-01-01

    ... 16 Commercial Practices 2 2010-01-01 2010-01-01 false Testing procedure for special packaging. 1700.20 Section 1700.20 Commercial Practices CONSUMER PRODUCT SAFETY COMMISSION POISON PREVENTION... their classroom or other supervised area by the tester or another adult. 29. If the children are...

  4. 40 CFR 205.57-4 - Testing procedures.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 24 2010-07-01 2010-07-01 false Testing procedures. 205.57-4 Section 205.57-4 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) NOISE ABATEMENT PROGRAMS TRANSPORTATION EQUIPMENT NOISE EMISSION CONTROLS Medium and Heavy Trucks § 205.57-4...

  5. 40 CFR 60.736 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 6 2010-07-01 2010-07-01 false Test methods and procedures. 60.736 Section 60.736 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS (CONTINUED) STANDARDS OF PERFORMANCE FOR NEW STATIONARY SOURCES Standards of Performance for Calciners...

  6. Test Procedure for Axially Loaded Piles in Sand

    DEFF Research Database (Denmark)

    Thomassen, Kristina

    The test procedure described in the following is used when examining the effects of static or cyclic loading on the skin friction of an axially loaded pile in dense sand. The pile specimen is only loaded in tension to avoid any contribution from the base resistance. The pile dimensions are chosen...

  7. AP-102/104 Retrieval control system qualification test procedure

    Energy Technology Data Exchange (ETDEWEB)

    RIECK, C.A.

    1999-05-18

    This Qualification Test Procedure documents the results of the qualification testing that was performed on the Project W-211, ''Initial Tank Retrieval Systems,'' retrieval control system (RCS) for tanks 241-AP-102 and 241-AP-104. The results confirm that the RCS has been programmed correctly and that the two related hardware enclosures have been assembled in accordance with the design documents.

  8. Pregnant women's attitudes for accepting or declining a serum-alpha-fetoprotein test.

    Science.gov (United States)

    Sandén, M L; Bjurulf, P

    1988-01-01

    This study analyzes attitudes which influence women in their decisions to accept or decline a serum-Alpha-FetoProtein (se-AFP) test. The survey covers all women who have visited antenatal clinics offering the se-AFP test in a given area during a specified period. All in all, the pregnant women answered four questionnaires: three before and one after delivery. 19% of the women declined the se-AFP test. During pregnancy, the women who did not take the se-AFP test said that the reason was anxiety about the results of the test. Many of the women who took the test said that they wanted assurance that the child was healthy. This entails obvious ethical problems when the test is positive. The test-routine does not have a major influence in the decision-making process. A change in routines to shorten the waiting time, and a different cut-off line might influence some of the declining women to accept the test.

  9. Interim qualification tests and procedures for terrestrial photovoltaic thin-film flat-plate modules

    Science.gov (United States)

    Deblasio, R.; Mrig, L.; Waddington, D.

    1990-01-01

    This document provides recommended procedures and specifications for qualification tests that are structured to evaluate terrestrial thin-film flat-plate photovoltaic nonconcentrating modules intended for power generation applications. The qualification tests provided in this document are designed to evaluate flat-plate thin-film photovoltaic (PV) module design performance and susceptibility to known failure mechanisms. Emphasis is placed on testing and evaluating module performance characteristics and design features that will affect possible degradation of module performance and physical properties resulting from solar exposure, environmental weathering, mechanical loading, corrosion, and module shadowing. Because of limited thin-film module field operation experience and the evolutionary nature of new thin-film module material technologies and designs, these tests should not be considered definitive or complete, nor do they provide a basis to predict 30-year field life. Current understanding of failure and degradation mechanisms and the relationship between accelerated tests and field reliability is not sufficient to allow accurate estimation of life-expectancy, nor are the cycling tests given in this document considered to be equivalent to a full 30-year field exposure. However, the test and evaluation procedures given in this document provide a common approach for conducting qualification tests. Acceptable results from these tests should provide reasonable assurance that the modules that pass these tests will perform reliably in the field but for an unspecified period of time.

  10. Preliminary evaluation of an analog procedure to assess acceptability of intimate partner violence against women: the Partner Violence Acceptability Movie Task

    OpenAIRE

    Gracia, Enrique; Rodriguez, Christina M.; Lila, Marisol

    2015-01-01

    Acceptability of partner violence against women is a risk factor linked to its perpetration, and to public, professionals’ and victims’ responses to this behavior. Research on the acceptability of violence in intimate partner relationships is, however, limited by reliance solely on self-reports that often provide distorted or socially desirable accounts that may misrepresent respondents’ attitudes. This study presents data on the development and initial validation of a new analog task assessi...

  11. Preliminary evaluation of an analog procedure to assess acceptability of intimate partner violence against women: The Partner Violence Acceptability Movie Task (PVAM)

    OpenAIRE

    Enrique eGracia; Rodriguez, Christina M.; Marisol eLila

    2015-01-01

    Acceptability of partner violence against women is a risk factor linked to its perpetration, and to public, professionals’ and victims’ responses to this behavior. Research on the acceptability of violence in intimate partner relationships is, however, limited by reliance solely on self-reports that often provide distorted or socially desirable accounts that may misrepresent respondents’ true attitudes. This study presents data on the development and initial validation of a new analog task as...

  12. Evaluations of the Optimal Discovery Procedure for Multiple Testing.

    Science.gov (United States)

    Rubin, Daniel B

    2016-05-01

    The Optimal Discovery Procedure (ODP) is a method for simultaneous hypothesis testing that attempts to gain power relative to more standard techniques by exploiting multivariate structure [1]. Specializing to the example of testing whether components of a Gaussian mean vector are zero, we compare the power of the ODP to a Bonferroni-style method and to the Benjamini-Hochberg method when the testing procedures aim to respectively control certain Type I error rate measures, such as the expected number of false positives or the false discovery rate. We show through theoretical results, numerical comparisons, and two microarray examples that when the rejection regions for the ODP test statistics are chosen such that the procedure is guaranteed to uniformly control a Type I error rate measure, the technique is generally less powerful than competing methods. We contrast and explain these results in light of previously proven optimality theory for the ODP. We also compare the ordering given by the ODP test statistics to the standard rankings based on sorting univariate p-values from smallest to largest. In the cases we considered the standard ordering was superior, and ODP rankings were adversely impacted by correlation. PMID:27227716

  13. 100 Area soil washing bench-scale test procedures

    International Nuclear Information System (INIS)

    This document describes methodologies and procedures for conducting soil washing treatability tests in accordance with the 100 Area Soil Washing Treatability Test Plan (DOE-RL 1992, Draft A). The objective of this treatability study is to evaluate the use of physical separation systems and chemical extraction methods as a means of separating chemically and radioactively contaminated soil fractions from uncontaminated soil fractions. These data will be primarily used for determining feasibility of the individual unit operations and defining the requirements for a system, or systems, for pilot-scale testing

  14. 100 Area soil washing bench-scale test procedures

    Energy Technology Data Exchange (ETDEWEB)

    Freeman, H.D.; Gerber, M.A.; Mattigod, S.V.; Serne, R.J.

    1993-03-01

    This document describes methodologies and procedures for conducting soil washing treatability tests in accordance with the 100 Area Soil Washing Treatability Test Plan (DOE-RL 1992, Draft A). The objective of this treatability study is to evaluate the use of physical separation systems and chemical extraction methods as a means of separating chemically and radioactively contaminated soil fractions from uncontaminated soil fractions. These data will be primarily used for determining feasibility of the individual unit operations and defining the requirements for a system, or systems, for pilot-scale testing.

  15. A Procedure for Controlling General Test Overlap in Computerized Adaptive Testing

    Science.gov (United States)

    Chen, Shu-Ying

    2010-01-01

    To date, exposure control procedures that are designed to control test overlap in computerized adaptive tests (CATs) are based on the assumption of item sharing between pairs of examinees. However, in practice, examinees may obtain test information from more than one previous test taker. This larger scope of information sharing needs to be…

  16. Mixed Variables-Attributes Test Plans for Single and DoubleAcceptance Sampling under Exponential Distribution

    OpenAIRE

    Dongdong Xiang; Xiaolong Pu; Yan Li

    2011-01-01

    The mixed variables-attributes test plans for single acceptance sampling are proposed to protect “good lots” from attributes aspect and to optimize sample sizes from variables aspect. For the single and double mixed plans, exact formulas of the operating characteristic and average sample number are developed for the exponential distribution. Numerical illustrations show that the mixed sampling plans have some advantages over the variables plans or attributes plans alone.

  17. Test report for run-in acceptance testing of Project W-151 300 HP mixing pumps

    Energy Technology Data Exchange (ETDEWEB)

    Berglin, B.G.

    1998-01-29

    This report documents the results of a performance demonstration and operational checkout of three 300 HP mixer pumps in accordance with WHC-SD-WI51-TS-001 ``Mixer Pump Test Specification for Project W-151`` and Statement of Work 8K520-EMN-95-004 ``Mixer Pump Performance Demonstration at MASF`` in the 400 Area Maintenance and Storage Facility (MASF) building. Testing of the pumps was performed by Fast Flux Test Facility (FFTF) Engineering and funded by the Tank Waste Remediation System (TWRS) Project W-151. Testing began with the first pump on 04-01-95 and ended with the third pump on 11-01-96. Prior to testing, the MASF was modified and prepared to meet the pump testing requirements set forth by the Test Specification and the Statement of Work.

  18. Using case-based reasoning for generating functional test procedures

    Directory of Open Access Journals (Sweden)

    Martha Dunia Delgado Dapena

    2010-10-01

    Full Text Available This paper presents a proposal for storage structure and retrieval mechanisms used for implementing case-based reasoning (CBR in generating functional test procedures in software projects. This proposal was based on software project t functional requirements and sets out the proposed algorithms for considering the similarity between each pair of projects as well as those leading to adapting the solution found in the case base.

  19. Cassini RTG Acceptance Test Results and RTG Performance on Galileo and Ulysses

    Science.gov (United States)

    Kelly, C. E.; Klee, P. M.

    1997-06-01

    Flight acceptance testing has been completed for the RTGs to be used on the Cassini spacecraft which is scheduled for an October 6, 1997 launch to Saturn. The acceptance test program includes vibration tests, magnetic field measurements, properties (weight and c.g.) and thermal vacuum test. This paper presents The thermal vacuum test results. Three RTGs are to be used, F 2, F 6, and F 7. F 5 is tile back up RTG, as it was for the Galileo and Ulysses missions launched in 1989 and 1990, respectively. RTG performance measured during the thermal vacuum tests carried out at die Mound Laboratory facility met all specification requirements. Beginning of mission (BOM) and end of mission (EOM) power predictions have been made based on than tests results. BOM power is predicted to be 888 watts compared to the minimum requirement of 826 watts. Degradation models predict the EOM power after 16 years is to be 640 watts compared to a minimum requirement of 596 watts. Results of small scale module tests are also showing. The modules contain couples from the qualification and flight production runs. The tests have exceeded 28,000 hours (3.2 years) and are continuing to provide increased confidence in the predicted long term performance of the Cassini RTGs. All test results indicate that the power requirements of the Cassini spacecraft will be met. BOM and EOM power margins of over five percent are predicted. Power output from telemetry for the two Galileo RTGs are shown from the 1989 launch to the recent Jupiter encounter. Comparisons of predicted, measured and required performance are shown. Telemetry data are also shown for the RTG on the Ulysses spacecraft which completed its planned mission in 1995 and is now in the extended mission.

  20. Cassini RTG acceptance test results and RTG performance on Galileo and Ulysses

    International Nuclear Information System (INIS)

    Flight acceptance testing has been completed for the RTGs to be used on the Cassini spacecraft which is scheduled for an October 6, 1997 launch to Saturn. The acceptance test program includes vibration tests, magnetic field measurements, mass properties (weight and c.g.) and thermal vacuum test. This paper presents the thermal vacuum test results. Three RTGs are to be used, F-2, F-6, and F-7. F-5 is the backup RTG, as it was for the Galileo and Ulysses missions launched in 1989 and 1990, respectively. RTG performance measured during the thermal vacuum tests carried out at the Mound Laboratory facility met all specification requirements. Beginning of mission (BOM) and end of mission (EOM) power predictions have been made based on these tests results. BOM power is predicted to be 888 watts compared to the minimum requirement of 826 watts. Degradation models predict the EOM power after 16 years is to be 640 watts compared to a minimum requirement of 596 watts. Results of small scale module tests are also shown. The modules contain couples from the qualification and flight production runs. The tests have exceeded 28,000 hours (3.2 years) and are continuing to provide increased confidence in the predicted long term performance of the Cassini RTGs. All test results indicate that the power requirements of the Cassini spacecraft will be met. BOM and EOM power margins of over 5% are predicted. Power output from telemetry for the two Galileo RTGs are shown from the 1989 launch to the recent Jupiter encounter. Comparisons of predicted, measured and required performance are shown. Telemetry data are also shown for the RTG on the Ulysses spacecraft which completed its planned mission in 1995 and is now in the extended mission

  1. Cassini RTG acceptance test results and RTG performance on Galileo and Ulysses

    International Nuclear Information System (INIS)

    Flight acceptance testing has been completed for the RTGs to be used on the Cassini spacecraft which is scheduled for an October 6, 1997 launch to Saturn. The acceptance test program includes vibration tests, magnetic field measurements, properties (weight and c.g.) and thermal vacuum test. This paper presents The thermal vacuum test results. Three RTGs are to be used, F-2, F-6, and F-7. F-5 is tile back-up RTG, as it was for the Galileo and Ulysses missions launched in 1989 and 1990, respectively. RTG performance measured during the thermal vacuum tests carried out at die Mound Laboratory facility met all specification requirements. Beginning of mission (BOM) and end of mission (EOM) power predictions have been made based on than tests results. BOM power is predicted to be 888 watts compared to the minimum requirement of 826 watts. Degradation models predict the EOM power after 16 years is to be 640 watts compared to a minimum requirement of 596 watts. Results of small scale module tests are also showing. The modules contain couples from the qualification and flight production runs. The tests have exceeded 28,000 hours (3.2 years) and are continuing to provide increased confidence in the predicted long term performance of the Cassini RTGs. All test results indicate that the power requirements of the Cassini spacecraft will be met. BOM and EOM power margins of over five percent are predicted. Power output from telemetry for the two Galileo RTGs are shown from the 1989 launch to the recent Jupiter encounter. Comparisons of predicted, measured and required performance are shown. Telemetry data are also shown for the RTG on the Ulysses spacecraft which completed its planned mission in 1995 and is now in the extended mission

  2. Quality assurance of radiopharmaceuticals - specifications and test procedures

    International Nuclear Information System (INIS)

    The authors report on a Radiopharmaceutical Quality Assurance Test Programme carried out by the Australian Radiation Laboratory in which radiopharmaceuticals used in nuclear medicine in Australia are tested for compliance with specifications. Where the radiopharmaceutical is the subject of a monograph in the British Pharmacopoeia or the European Pharmacopoeia, then the specifications given in the Pharmacopoeia are adopted. In other cases the specifications given have been adopted by this Laboratory and have no legal status. In some cases test procedures described have been taken from various Pharmacopoeias or methods published in the literature. In other cases test methods described have been developed at this Laboratory. It should be noted that, unless stated otherwise, specifications listed apply at all times up until product expire

  3. Study on service test procedure for SMES systems

    Science.gov (United States)

    Ren, L.; Tang, Y.; Li, J.; Shi, J.; Hu, Y.; Guo, F.; Chen, L.; Cheng, S.

    2009-10-01

    Some superconducting magnetic energy storage (SMES) systems have been developed successfully, and the function of them was verified by experiment. Considering the impact on the system security and stability when they are connected to the power system, it is important to build a standard test procedure and propose the main experimental contents for some representative superconducting devices. Based on the experimental results of a 35 kJ/7 kW conduction-cooled high- T c SMES system, which was developed by Huazhong University of Science and Technology, the test items and methods for a conduction-cooled high- T c SMES system are studied. For performance evaluation of this high- T c SMES, three superconducting magnet performance tests, power conditioning characteristic test and dynamic experiment in a simulated power system were fulfilled.

  4. Study on service test procedure for SMES systems

    Energy Technology Data Exchange (ETDEWEB)

    Ren, L., E-mail: rl8868@126.co [R and D Center of Applied Superconductivity, Huazhong University of Science and Technology, Wuhan 430074 (China); Tang, Y.; Li, J.; Shi, J. [R and D Center of Applied Superconductivity, Huazhong University of Science and Technology, Wuhan 430074 (China); Hu, Y. [State Grid Electric Power Research Institute, Wuhan 430074 (China); Guo, F.; Chen, L.; Cheng, S. [R and D Center of Applied Superconductivity, Huazhong University of Science and Technology, Wuhan 430074 (China)

    2009-10-15

    Some superconducting magnetic energy storage (SMES) systems have been developed successfully, and the function of them was verified by experiment. Considering the impact on the system security and stability when they are connected to the power system, it is important to build a standard test procedure and propose the main experimental contents for some representative superconducting devices. Based on the experimental results of a 35 kJ/7 kW conduction-cooled high-T{sub c} SMES system, which was developed by Huazhong University of Science and Technology, the test items and methods for a conduction-cooled high-T{sub c} SMES system are studied. For performance evaluation of this high-T{sub c} SMES, three superconducting magnet performance tests, power conditioning characteristic test and dynamic experiment in a simulated power system were fulfilled.

  5. Space Suit Portable Life Support System (PLSS) 2.0 Pre-Installation Acceptance (PIA) Testing

    Science.gov (United States)

    Anchondo, Ian; Cox, Marlon; Meginnis, Carly; Westheimer, David; Vogel, Matt R.

    2016-01-01

    Following successful completion of the space suit Portable Life Support System (PLSS) 1.0 development and testing in 2011, the second system-level prototype, PLSS 2.0, was developed in 2012 to continue the maturation of the advanced PLSS design. This advanced PLSS is intended to reduce consumables, improve reliability and robustness, and incorporate additional sensing and functional capabilities over the current Space Shuttle/International Space Station Extravehicular Mobility Unit (EMU) PLSS. PLSS 2.0 represents the first attempt at a packaged design comprising first generation or later component prototypes and medium fidelity interfaces within a flight-like representative volume. Pre-Installation Acceptance (PIA) is carryover terminology from the Space Shuttle Program referring to the series of test sequences used to verify functionality of the EMU PLSS prior to installation into the Space Shuttle airlock for launch. As applied to the PLSS 2.0 development and testing effort, PIA testing designated the series of 27 independent test sequences devised to verify component and subsystem functionality, perform in situ instrument calibrations, generate mapping data, define set-points, evaluate control algorithms, evaluate hardware performance against advanced PLSS design requirements, and provide quantitative and qualitative feedback on evolving design requirements and performance specifications. PLSS 2.0 PIA testing was carried out in 2013 and 2014 using a variety of test configurations to perform test sequences that ranged from stand-alone component testing to system-level testing, with evaluations becoming increasingly integrated as the test series progressed. Each of the 27 test sequences was vetted independently, with verification of basic functionality required before completion. Because PLSS 2.0 design requirements were evolving concurrently with PLSS 2.0 PIA testing, the requirements were used as guidelines to assess performance during the tests; after the

  6. Acceptance sampling plan of accelerated life testing for lognormal distribution under time-censoring

    Directory of Open Access Journals (Sweden)

    Li Xiaoyang

    2015-06-01

    Full Text Available Lognormal distribution is commonly used in engineering. It is also a life distribution of important research values. For long-life products follow this distribution, it is necessary to apply accelerated testing techniques to product demonstration. This paper describes the development of accelerated life testing sampling plans (ALSPs for lognormal distribution under time-censoring conditions. ALSPs take both producer and consumer risks into account, and they can be designed to work whether acceleration factor (AF is known or unknown. When AF is known, life testing is assumed to be conducted under accelerated conditions with time-censoring. The producer and consumer risks are satisfied, and the size of test sample and the size of acceptance number are optimized. Then sensitivity analyses are conducted. When AF is unknown, two or more predetermined levels of accelerated stress are used. The sample sizes and sample proportion allocated to each stress level are optimized. The acceptance constant that satisfies producer and consumer risk is obtained by minimizing the generalized asymptotic variance of the test statistics. Finally, the properties of the two ALSPs (one for known-AF conditions and one for unknown-AF conditions are investigated to show that the proposed method is correct and usable through numerical examples.

  7. Consumer acceptance, market test and market development of irradiated rice, dehydrated vegetables and spices

    International Nuclear Information System (INIS)

    Establishment of irradiation processing parameters, a quality assurance system, consumer acceptance, market test and market development of irradiated rice, dehydrated vegetables and spices were the activities carried out in this project by the Chinese Agricultural Irradiation Center. The results of the studies showed that the process dose for rice was 0.2-0.5 kGy when the non-uniformity was lower than 2.5, dose range for dehydrated vegetables was 5-7 kGy, dose for spices was 7-8 kGy. The system for quality assurance was established. The processing standards for several irradiated food items were set up. Market test showed that more than 70-80% of consumers accepted irradiated food. Industrial companies also accepted irradiated dehydrated vegetables and spices. The latter were successfully introduced to the markets and successful commercialization of irradiated garlic was followed. The economic benefit of operating the Chinese Agricultural Irradiation Center was analyzed and found attractive, especially for low dose irradiation of foods in sufficient supply. (author)

  8. User Acceptance Test of Computer-Assisted Problem-Based Learning Assessment Tool (CAPBLAT

    Directory of Open Access Journals (Sweden)

    Muhammad Qomaruddin

    2014-02-01

    Full Text Available Problem-based Learning (PBL is an instructional design method that is used by many lecturers to create more efficient and meaningful learning experiences. PBL emphasizes the lecturer in facilitating in collaborative learning. They conventionally play a little role in the formal assessment process. Peer-assessment and self-assessment are the most frequent methods of assessment employed by the lecturers in conducting PBL approach. This paper presents acceptance testing of a computer-based tool for peer-assessment and self-assessment in PBL approach, which called Computer-Assisted Problem-Based Learning Assessment Tool (CAPBLAT. The tool was designed to assist a lecturer in conducting PBL teaching method and assess students learning progress. Two PBL classes participated in the testing of the tool. During the study, questionnaires were administered to students. The results concerning the assessment tool acceptance demonstrate that using CAPBLAT in the PBL assessment process received better acceptance from both the students and lecturers.

  9. Geoscience Laser Altimeter System (GLAS) Instrument: Flight Loop Heat Pipe (LHP) Acceptance Thermal Vacuum Test

    Science.gov (United States)

    Baker, Charles; Butler, Dan; Ku, Jentung; Grob, Eric; Swanson, Ted; Nikitkin, Michael; Powers, Edward I. (Technical Monitor)

    2001-01-01

    Two loop heat pipes (LHPs) are to be used for tight thermal control of the Geoscience Laser Altimeter System (GLAS) instrument, planned for flight in late 2001. The LHPs are charged with Propylene as a working fluid. One LHP will be used to transport 110 W from a laser to a radiator, the other will transport 160 W from electronic boxes to a separate radiator. The application includes a large amount of thermal mass in each LHP system and low initial startup powers. The initial design had some non-ideal flight design compromises, resulted in a less than ideal charge level for this design concept with a symmetrical secondary wick. This less than ideal charge was identified as the source of inadequate performance of the flight LHPs during the flight thermal vacuum test in October of 2000. We modified the compensation chamber design, re-built and charged the LHPs for a final LHP acceptance thermal vacuum test. This test performed March of 2001 was 100% successful. This is the last testing to be performed on the LHPs prior to instrument thermal vacuum test. This sensitivity to charge level was shown through varying the charge on a Development Model Loop Heat Pipe (DM LHP) and evaluating performance at various fill levels. At lower fills similar to the original charge in the flight units, the same poor performance was observed. When the flight units were re-designed and filled to the levels similar to the initial successful DM LHP test, the flight units also successfully fulfilled all requirements. This final flight Acceptance test assessed performance with respect to startup, low power operation, conductance, and control heater power, and steady state control. The results of the testing showed that both LHPs operated within specification. Startup on one of the LHPs was better than the other LHP because of the starter heater placement and a difference in evaporator design. These differences resulted in a variation in the achieved superheat prior to startup. The LHP with

  10. Perceived acceptability of home-based couples voluntary HIV counseling and testing in Northern Tanzania.

    Science.gov (United States)

    Njau, B; Watt, M H; Ostermann, J; Manongi, R; Sikkema, K J

    2012-01-01

    It is estimated that 5.6% of the Tanzanian population ages 15-49 are infected with HIV, but only 30% of adults have ever had an HIV test. Couples' testing has proven to increase testing coverage and introduce HIV prevention, but barriers include access to testing services and unequal gender dynamics in relationships. Innovative approaches are needed to address barriers to couple's testing and increase uptake of HIV testing. Using qualitative data collection methods, a formative study was conducted to assess the acceptability of a home-based couples counseling and testing (HBCCT) approach. Eligible study participants included married men and women, HIV-infected individuals, health care and home-based care providers, voluntary counseling and testing counselors, and community leaders. A total of 91 individuals participated in focus group discussions (FGDs) and in-depth interviews conducted between September 2009 and January 2010 in rural settings in Northern Tanzania. An HBCCT intervention appears to be broadly acceptable among participants. Benefits of HBCCT were identified in terms of access, confidentiality, and strengthening the relationship. Fears of negative consequences from knowing one's HIV status, including stigma, blame, physical abuse, or divorce, remain a concern and a potential barrier to the successful provision of the intervention. Lessons for implementation highlighted the importance of appointments for home visits, building relationships of confidence and trust between counselors and clients, and assessing and responding to a couple's readiness to undergo HIV testing. HBCCT should addresses HIV stigma, emphasize confidentiality, and improve communication skills for disclosure and decision-making among couples.

  11. Home-based voluntary HIV counselling and testing found highly acceptable and to reduce inequalities

    Directory of Open Access Journals (Sweden)

    Michelo Charles

    2010-06-01

    Full Text Available Abstract Background Low uptake of voluntary HIV counselling and testing (VCT in sub-Saharan Africa is raising acceptability concerns which might be associated with ways by which it is offered. We investigated the acceptability of home-based delivery of counselling and HIV testing in urban and rural populations in Zambia where VCT has been offered mostly from local clinics. Methods A population-based HIV survey was conducted in selected communities in 2003 (n = 5035. All participants stating willingness to be HIV tested were offered VCT at home and all counselling was conducted in the participants' homes. In the urban area post-test counselling and giving of results were done the following day whereas in rural areas this could take 1-3 weeks. Results Of those who indicated willingness to be HIV tested, 76.1% (95%CI 74.9-77.2 were counselled and received the test result. Overall, there was an increase in the proportion ever HIV tested from 18% before provision of home-based VCT to 38% after. The highest increase was in rural areas; among young rural men aged 15-24 years up from 14% to 42% vs. for urban men from 17% to 37%. Test rates by educational attainment changed from being positively associated to be evenly distributed after home-based VCT. Conclusions A high uptake was achieved by delivering HIV counselling and testing at home. The highest uptakes were seen in rural areas, in young people and groups with low educational attainment, resulting in substantial reductions in existing inequalities in accessing VCT services.

  12. Evaluation of WES one-dimensional dynamic soil testing procedures

    Science.gov (United States)

    Seaman, L.

    1983-06-01

    The Geomechanics Division of the Waterways Experiment Station (WES) tests soils to provide material property data for ground motion calculations in support of high explosives experiments. Thus, the validity of the ground motion predictions depends greatly on the WES material properties used. The usual WES experiments are uniaxial strain (UX) and triaxial shear (TX) tests that can be performed at static testing rates and at dynamic rates with rise times down to about one millisecond. However, in the field experiments being simulated, the measured rise times are often 0.01 ms up to 0.1 ms. Recently, an effort has been made to obtain test data in UX and TX devices with rise times of about 0.1 ms. Some of the test data obtained at these very high rates have appeared to be anomalous, suggesting that wave propagation or other effects may be invalidating the data. Because of these questions about the high rate soils test data, WES asked SRI to evaluate the testing methods used at WES and the data reduction device. This report documents an evaluation of laboratory test data obtained from the explosive-loaded uniaxial strain device developed at WES and the data analysis procedures currently being used.

  13. Project W-314 acceptance test report HNF-4643 for HNF-4642 241-AN-A valve pit manifold valves and position indication for project W-314

    Energy Technology Data Exchange (ETDEWEB)

    HAMMERS, J.S.

    1999-09-22

    The purpose of the test was to verify that the AN Tank Farm Manifold Valves can be manually manipulated to the required operating position and that the electrical and visual indications accurately reflect that position. Physical locking devices were also verified to function. The Acceptance Test Procedure HNF-4642, 241-AN-A Valve Pit Manifold Valves and Position Indication was conducted between 23 June and 10 August 1999 at the 200E AN Tank Farm. The test has no open test exceptions. The test was conducted prior to final engineering ''as built'' activities being completed, this had an impact on the procedure and test results, ECN 653752 was written to correct the mismatch between the procedure and actual field conditions. P&ID H-14-100941 was changed via ECN-W-314-4C-120. All components, identified in the procedure, were not found to be labeled and identified as written in the procedure, temporary tags were used for operational identification. A retest of valve ANA-WT-V 318 was required because it was removed from its installed position and modified after testing was completed.

  14. Project W-314 acceptance test report HNF-4643 for HNF-4642 241-AN-A valve pit manifold valves and position indication for project W-314

    International Nuclear Information System (INIS)

    The purpose of the test was to verify that the AN Tank Farm Manifold Valves can be manually manipulated to the required operating position and that the electrical and visual indications accurately reflect that position. Physical locking devices were also verified to function. The Acceptance Test Procedure HNF-4642, 241-AN-A Valve Pit Manifold Valves and Position Indication was conducted between 23 June and 10 August 1999 at the 200E AN Tank Farm. The test has no open test exceptions. The test was conducted prior to final engineering ''as built'' activities being completed, this had an impact on the procedure and test results, ECN 653752 was written to correct the mismatch between the procedure and actual field conditions. P and ID H-14-100941 was changed via ECN-W-314-4C-120. All components, identified in the procedure, were not found to be labeled and identified as written in the procedure, temporary tags were used for operational identification. A retest of valve ANA-WT-V 318 was required because it was removed from its installed position and modified after testing was completed

  15. Lightning arrestor connector lead magnesium niobate qualification pellet test procedures.

    Energy Technology Data Exchange (ETDEWEB)

    Tuohig, W. (Honeywell FM& T, Kansas City, MO); Mahoney, Patrick A.; Tuttle, Bruce Andrew; Wheeler, Jill Susanne

    2009-02-01

    Enhanced knowledge preservation for DOE DP technical component activities has recently received much attention. As part of this recent knowledge preservation effort, improved documentation of the sample preparation and electrical testing procedures for lead magnesium niobate--lead titanate (PMN/PT) qualification pellets was completed. The qualification pellets are fabricated from the same parent powders used to produce PMN/PT lightning arrestor connector (LAC) granules at HWF&T. In our report, the procedures for fired pellet surface preparation, electrode deposition, electrical testing and data recording are described. The dielectric measurements described in our report are an information only test. Technical reasons for selecting the electrode material, electrode size and geometry are presented. The electrical testing is based on measuring the dielectric constant and dissipation factor of the pellet during cooling from 280 C to 220 C. The most important data are the temperature for which the peak dielectric constant occurs (Curie Point temperature) and the peak dielectric constant magnitude. We determined that the peak dielectric constant for our procedure would be that measured at 1 kHz at the Curie Point. Both the peak dielectric constant and the Curie point parameters provide semi-quantitative information concerning the chemical and microstructural homogeneity of the parent material used for the production of PMN/PT granules for LACs. Finally, we have proposed flag limits for the dielectric data for the pellets. Specifically, if the temperature of the peak dielectric constant falls outside the range of 250 C {+-} 30 C we propose that a flag limit be imposed that will initiate communication between production agency and design agency personnel. If the peak dielectric constant measured falls outside the range 25,000 {+-} 10,000 we also propose that a flag limit be imposed.

  16. 21 CFR 355.70 - Testing procedures for fluoride dentifrice drug products.

    Science.gov (United States)

    2010-04-01

    ... 21 Food and Drugs 5 2010-04-01 2010-04-01 false Testing procedures for fluoride dentifrice drug... Procedures § 355.70 Testing procedures for fluoride dentifrice drug products. (a) A fluoride dentifrice drug... tests: Enamel solubility reduction or fluoride enamel uptake. The testing procedures for...

  17. Neurophysiological tests and neuroimaging procedures in non-acute headache

    DEFF Research Database (Denmark)

    Sandrini, G; Friberg, L; Jänig, W;

    2004-01-01

    history suggests a possible diagnosis of epilepsy (differential diagnosis). Ictal EEG could be useful in certain patients suffering from hemiplegic and basilar migraine. 2 Recording of evoked potentials is not recommended for the diagnosis of headache disorders. 3 There is no evidence to justify......The use of instrumental examinations in headache patients varies widely. In order to evaluate their usefulness, the most common instrumental procedures were evaluated, on the basis of evidence from the literature, by an EFNS Task Force (TF) on neurophysiological tests and imaging procedures in non......-acute headache patients. The conclusions of the TF regarding each technique are expressed in the following guidelines for clinical use. 1 Interictal electroencephalography (EEG) is not routinely indicated in the diagnostic evaluation of headache patients. Interictal EEG is, however, indicated if the clinical...

  18. Ten years experience in acceptance testing of diagnostic radiology equipment in Hungary

    International Nuclear Information System (INIS)

    The EURATOM directive for the radiation protection of patients, which declares the mandatory evaluation of the performance of diagnostic X-ray machines, took effect in 1997. In Hungary, this became a harmonized act, the 31/2001. (X. 3.) decree of the Minister of Health. The most highlighted segment of the evaluation of technical performance is acceptance testing. The NRIRR Department of Radiohygiene performs acceptance tests since 2002. From then on we have performed more than 400 examinations on different modalities such as radiographic, fluoroscopic, DSA, mammographic, CT and dental panoramic X-rays. The tests were performed according to IEC = EN = MSZ EN international, European and at the same time Hungarian standards. This presentation discusses our experience and gives a comprehensive overview of the examinations by the evaluation of the measurement data. Our results show the importance of testing, done independently from the manufacturer and installer: in est. 25 % of the cases overall we have found inadequacies which required remedial action on the part of the service

  19. The Rapid Integration and Test Environment - A Process for Achieving Software Test Acceptance

    OpenAIRE

    Rick Jack

    2010-01-01

    The Rapid Integration and Test Environment (RITE) initiative, implemented by the Program Executive Office, Command, Control, Communications, Computers and Intelligence, Command and Control Program Office (PMW-150), was born of necessity. Existing processes for requirements definition and management, as well as those for software development, did not consistently deliver high-quality Navy command and control (C2) systems on time and within budget. Navy C2 software programs experienced an inc...

  20. Operational test procedure for Bldg 241-A-701 air compressor

    International Nuclear Information System (INIS)

    This document is an Operability Test Procedure (OTP) which will verify and record that the 241-A-701 air compressor and associated equipment operates within their intended design parameters. The activities defined in this OTP will be performed to ensure the daily operation of the new compressed air system can be reliable and efficient. The Compressed Air System (CAS) for 241-A-701 supplies process and instrument air to the A, AX, AY, and AZ tank farms. The primary use of the CAS is for tank farms instrumentation, air operated valves, and air lift circulators

  1. Multi-Canister Overpack (MCO) Combustible Gas Management Leak Test Acceptance Criteria (OCRWM)

    International Nuclear Information System (INIS)

    The purpose of this document is to support the Spent Nuclear Fuel Project's combustible gas management strategy while avoiding the need to impose any requirements for oxygen free atmospheres within storage tubes that contain multi-canister overpacks (MCO). In order to avoid inerting requirements it is necessary to establish and confirm leak test acceptance criteria for mechanically sealed and weld sealed MCOs that are adequte to ensure that, in the unlikely event the leak test results for any MCO were to approach either of those criteria, it could still be handled and stored in stagnant air without compromising the SNF Project's overall strategy to prevent accumulation of combustible gas mixtures within MCOs or within their surroundings. To support that strategy, this document: (1) establishes combustible gas management functions and minimum functional requirements for the MCO's mechanical seals and closure weld(s); (2) establishes a maximum practical value for the minimum required initial MCO inert backfill gas pressure; and (3) based on items 1 and 2, establishes and confirms leak test acceptance criteria for the MCO's mechanical seal and final closure weld(s)

  2. Multi Canister Overpack (MCO) Combustible Gas Management Leak Test Acceptance Criteria (OCRWM)

    Energy Technology Data Exchange (ETDEWEB)

    SHERRELL, D.L.

    2000-10-10

    The purpose of this document is to support the Spent Nuclear Fuel Project's combustible gas management strategy while avoiding the need to impose any requirements for oxygen free atmospheres within storage tubes that contain multi-canister overpacks (MCO). In order to avoid inerting requirements it is necessary to establish and confirm leak test acceptance criteria for mechanically sealed and weld sealed MCOs that are adequte to ensure that, in the unlikely event the leak test results for any MCO were to approach either of those criteria, it could still be handled and stored in stagnant air without compromising the SNF Project's overall strategy to prevent accumulation of combustible gas mixtures within MCOs or within their surroundings. To support that strategy, this document: (1) establishes combustible gas management functions and minimum functional requirements for the MCO's mechanical seals and closure weld(s); (2) establishes a maximum practical value for the minimum required initial MCO inert backfill gas pressure; and (3) based on items 1 and 2, establishes and confirms leak test acceptance criteria for the MCO's mechanical seal and final closure weld(s).

  3. Accuracy and user-acceptability of HIV self-testing using an oral fluid-based HIV rapid test.

    Directory of Open Access Journals (Sweden)

    Oon Tek Ng

    Full Text Available BACKGROUND: The United States FDA approved an over-the-counter HIV self-test, to facilitate increased HIV testing and earlier linkage to care. We assessed the accuracy of self-testing by untrained participants compared to healthcare worker (HCW testing, participants' ability to interpret sample results and user-acceptability of self-tests in Singapore. METHODOLOGY/PRINCIPAL FINDINGS: A cross-sectional study, involving 200 known HIV-positive patients and 794 unknown HIV status at-risk participants was conducted. Participants (all without prior self-test experience performed self-testing guided solely by visual instructions, followed by HCW testing, both using the OraQuick ADVANCE Rapid HIV 1/2 Antibody Test, with both results interpreted by the HCW. To assess ability to interpret results, participants were provided 3 sample results (positive, negative, and invalid to interpret. Of 192 participants who tested positive on HCW testing, self-testing was positive in 186 (96.9%, negative in 5 (2.6%, and invalid in 1 (0.5%. Of 794 participants who tested negative on HCW testing, self-testing was negative in 791 (99.6%, positive in 1 (0.1%, and invalid in 2 (0.3%. Excluding invalid tests, self-testing had sensitivity of 97.4% (95% CI 95.1% to 99.7% and specificity of 99.9% (95% CI: 99.6% to 100%. When interpreting results, 96%, 93.1% and 95.2% correctly read the positive, negative and invalid respectively. There were no significant demographic predictors for false negative self-testing or wrongly interpreting positive or invalid sample results as negative. Eighty-seven percent would purchase the kit over-the-counter; 89% preferred to take HIV tests in private. 72.5% and 74.9% felt the need for pre- and post-test counseling respectively. Only 28% would pay at least USD15 for the test. CONCLUSIONS/SIGNIFICANCE: Self-testing was associated with high specificity, and a small but significant number of false negatives. Incorrectly identifying model results as

  4. 40 CFR 86.165-12 - Air conditioning idle test procedure.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 18 2010-07-01 2010-07-01 false Air conditioning idle test procedure... Complete Heavy-Duty Vehicles; Test Procedures § 86.165-12 Air conditioning idle test procedure. (a) Applicability. This section describes procedures for determining air conditioning-related CO2 emissions...

  5. EVALUATION OF ARG-1 SAMPLES PREPARED BY CESIUM CARBONATE DISSOLUTION DURING THE ISOLOK SME ACCEPTABILITY TESTING

    Energy Technology Data Exchange (ETDEWEB)

    Edwards, T.; Hera, K.; Coleman, C.

    2011-12-05

    Evaluation of Defense Waste Processing Facility (DWPF) Chemical Process Cell (CPC) cycle time identified several opportunities to improve the CPC processing time. The Mechanical Systems & Custom Equipment Development (MS&CED) Section of the Savannah River National Laboratory (SRNL) recently completed the evaluation of one of these opportunities - the possibility of using an Isolok sampling valve as an alternative to the Hydragard valve for taking DWPF process samples at the Slurry Mix Evaporator (SME). The use of an Isolok for SME sampling has the potential to improve operability, reduce maintenance time, and decrease CPC cycle time. The SME acceptability testing for the Isolok was requested in Task Technical Request (TTR) HLW-DWPF-TTR-2010-0036 and was conducted as outlined in Task Technical and Quality Assurance Plan (TTQAP) SRNLRP-2011-00145. RW-0333P QA requirements applied to the task, and the results from the investigation were documented in SRNL-STI-2011-00693. Measurement of the chemical composition of study samples was a critical component of the SME acceptability testing of the Isolok. A sampling and analytical plan supported the investigation with the analytical plan directing that the study samples be prepared by a cesium carbonate (Cs{sub 2}CO{sub 3}) fusion dissolution method and analyzed by Inductively Coupled Plasma - Optical Emission Spectroscopy (ICP-OES). The use of the cesium carbonate preparation method for the Isolok testing provided an opportunity for an additional assessment of this dissolution method, which is being investigated as a potential replacement for the two methods (i.e., sodium peroxide fusion and mixed acid dissolution) that have been used at the DWPF for the analysis of SME samples. Earlier testing of the Cs{sub 2}CO{sub 3} method yielded promising results which led to a TTR from Savannah River Remediation, LLC (SRR) to SRNL for additional support and an associated TTQAP to direct the SRNL efforts. A technical report resulting

  6. Recommended procedures for performance testing of radiobioassay laboratories: Volume 3, In vivo test phantoms

    International Nuclear Information System (INIS)

    Draft American National Standards Institute (ANSI) Standard N13.30 (Performance Criteria for Radiobioassay) was developed for the US Department of Energy and the US Nuclear Regulatory Commission to help ensure that bioassay laboratories provide accurate and consistent results. The draft standard describes the procedures necessary to establish a bioassay performance-testing laboratory and program. The bioassay performance-testing laboratory will conduct tests to evaluate the performance of service laboratories. Pacific Northwest Laboratory helped develop testing procedures as part of an effort to evaluate the draft ANSI N13.30 performance criteria by testing the existing measurement capabilities of various bioassay laboratories. This report recommends guidelines for the preparation, handling, storage, distribution, shipping, and documentation of test phantoms used for calibration of measurement systems for direct bioassay. The data base and recommended records system for documenting radiobioassay performance at the service laboratories are also presented

  7. Project W-314 performance mock-up test procedure

    Energy Technology Data Exchange (ETDEWEB)

    CARRATT, R.T.

    1999-06-24

    The purpose of this Procedure is to assist construction in the pre-operational fabrication and testing of the pit leak detection system and the low point drain assembly by: (1) Control system testing of the pit leak detection system will be accomplished by actuating control switches and verifying that the control signal is initiated, liquid testing and overall operational requirements stated in HNF-SD-W314-PDS-003, ''Project Development Specification for Pit Leak Detection''. (2) Testing of the low point floor drain assembly by opening and closing the drain to and from the ''retracted'' and ''sealed'' positions. Successful operation of this drain will be to verify that the seal does not leak on the ''sealed'' position, the assembly holds liquid until the leak detector actuates and the assembly will operate from on top of the mock-up cover block.

  8. 10 CFR 26.91 - Acceptable devices for conducting initial and confirmatory tests for alcohol and methods of use.

    Science.gov (United States)

    2010-01-01

    ... or that has passed its expiration date. (3) For ASDs that test breath specimens and meet EBT.... Alcohol screening devices (ASDs), including devices that test specimens of oral fluids or breath, must be...) Acceptable evidential breath testing devices. Evidential breath testing devices listed in the NHTSA CPL...

  9. 78 FR 45010 - Modifications to the Disability Determination Procedures; Extension of Testing of Some Disability...

    Science.gov (United States)

    2013-07-25

    ... procedures for adjudicating claims for disability insurance benefits under title II of the Social Security... ADMINISTRATION Modifications to the Disability Determination Procedures; Extension of Testing of Some Disability... involving modifications to the disability determination procedures. SUMMARY: We are announcing the...

  10. 77 FR 35464 - Modifications to the Disability Determination Procedures; Extension of Testing of Some Disability...

    Science.gov (United States)

    2012-06-13

    ... procedures for adjudicating claims for disability insurance benefits under title II of the Social Security... ADMINISTRATION Modifications to the Disability Determination Procedures; Extension of Testing of Some Disability... Involving Modifications to the Disability Determination Procedures. SUMMARY: We are announcing the...

  11. Embedding Procedure Assistance into Mission Control Tools Project

    Data.gov (United States)

    National Aeronautics and Space Administration — Procedures are the accepted means of commanding spacecraft. Procedures encode the operational knowledge of a system as derived from system experts, testing,...

  12. Test plan/procedure for the checkout of the USA cable communications test configuration for the electromagnetic compatibility (EMC) tests

    Science.gov (United States)

    Perry, J. C.

    1975-01-01

    A series of electromagnetic compatibility (EMC) tests were conducted in May, 1975 in the Soviet Union. The purpose of the EMC tests was to determine the effects of the operating environment of the Soviet aircraft, Soyuz, upon the electrical performance of the USA's cable communications equipment located in Soyuz. The test procedures necessary to check out the cable communications test configuration in preparation for the EMC tests are presented.

  13. 78 FR 5075 - Energy Conservation Program: Test Procedure for Set-Top Boxes

    Science.gov (United States)

    2013-01-23

    ... for determining the power consumption of a STB in the on, sleep, and off modes. The proposed test... the STB in sleep mode, DOE developed the test procedure from the sleep mode test procedure specified... test. For the sleep mode test, DOE proposes two tests: (1) a manual sleep test in which the STB...

  14. Refractory Metal Heat Pipe Life Test - Test Plan and Standard Operating Procedures

    Science.gov (United States)

    Martin, J. J.; Reid, R. S.

    2010-01-01

    Refractory metal heat pipes developed during this project shall be subjected to various operating conditions to evaluate life-limiting corrosion factors. To accomplish this objective, various parameters shall be investigated, including the effect of temperature and mass fluence on long-term corrosion rate. The test series will begin with a performance test of one module to evaluate its performance and to establish the temperature and power settings for the remaining modules. The performance test will be followed by round-the-clock testing of 16 heat pipes. All heat pipes shall be nondestructively inspected at 6-month intervals. At longer intervals, specific modules will be destructively evaluated. Both the nondestructive and destructive evaluations shall be coordinated with Los Alamos National Laboratory. During the processing, setup, and testing of the heat pipes, standard operating procedures shall be developed. Initial procedures are listed here and, as hardware is developed, will be updated, incorporating findings and lessons learned.

  15. Procedural development for nuclear waste canister impact testing

    International Nuclear Information System (INIS)

    Double containment requirements for transporting nuclear waste in glass form are costly and may not be necessary for some waste forms. To allow single containment, a procedure for examining particle size distribution and the amount of respirable particles generated under accident conditions was needed. A statistically designed experiment was conducted to examine the effects of glass temperature, fill rate and canister drop orientation upon the amount of sub-ten micron particles generated under simulated accident conditions. Measuring such small particles is somewhat inaccurate because of material loss in handling. By assuming a lognormal particle size distribution, the amount of sub-ten micron particles was estimated from the results for the larger measurable particles. Analyses revealed no temperature or fill rate effect but indicated that the amount of respirable particles is affected by drop orientation. This led to identification of a worst case drop orientation to be used in qualification testing. 4 refs., 2 figs

  16. Acceptance test for 900 MWe PWR unit replacement steam generators; Essai de reception des generateurs de vapeur de remplacement des tranches REP 900

    Energy Technology Data Exchange (ETDEWEB)

    Gourguechon, B.

    1993-12-31

    During the first half of 1994, the Gravelines 1 steam generators will be replaced (SG replacement procedure). The new SG`s differ from the former components notably by the alloy used for the tube bundle, in this case, the high chromium content Inconel 690. So, from this standpoint, they are to be considered as PWR 900 replacement SG first models and their thermal efficiency has consequently to be assessed. This will provide an opportunity of ensuring that the performance of the components delivered is in compliance with requirements and of making the necessary provisions if significant deviations are observed. The EFMT branch, which has been in charge of the instrumentation and acceptance of the different SG first models since the first PWR plants were commissioned, will be responsible for the acceptance tests and the ultimate validation of a performance assessment procedure applicable to the future replacement steam generators. The methods and tests proposed for SG expert appraisal are based on consideration of the importance of primary measurement quality for satisfactory SG assessment and of the new test facilities with which the 900 and 1 300 PWR plants are gradually being equipped. These facilities provide an on-site computer environment for tests compatible with the tools (PATTERN, etc.) used at EFMT and in other departments. This test is the first of this kind performed by EFMT and the test facility of a nuclear power plant. (author). 6 figs.

  17. Understanding Student Teachers’ Behavioural Intention to Use Technology: Technology Acceptance Model (TAM Validation and Testing

    Directory of Open Access Journals (Sweden)

    Kung-Teck, Wong

    2013-01-01

    Full Text Available This study sets out to validate and test the Technology Acceptance Model (TAM in the context of Malaysian student teachers’ integration of their technology in teaching and learning. To establish factorial validity, data collected from 302 respondents were tested against the TAM using confirmatory factor analysis (CFA, and structural equation modelling (SEM was used for model comparison and hypotheses testing. The goodness-of-fit test of the analysis shows partial support of the applicability of the TAM in a Malaysian context. Overall, the TAM accounted for 37.3% of the variance in intention to use technology among student teachers and of the five hypotheses formulated, four are supported. Perceived usefulness is a significant influence on attitude towards computer use and behavioural intention. Perceived ease of use significantly influences perceived usefulness, and finally, behavioural intention is found to be influenced by attitude towards computer use. The findings of this research contribute to the literature by validating the TAM in the Malaysian context and provide several prominent implications for the research and practice of technology integration development.

  18. An Application of Graphical Approach to Construct Multiple Testing Procedure in a Hypothetical Phase III Design

    Directory of Open Access Journals (Sweden)

    Naitee eTing

    2014-01-01

    Full Text Available Many multiple testing procedures (MTP have been developed in recent years. Among these new procedures, the graphical approach is flexible and easy to communicate with non-statisticians. A hypothetical Phase III clinical trial design is introduced in this manuscript to demonstrate how graphical approach can be applied in clinical product development. In this design, an active comparator is used. It is thought that this test drug under development could potentially be superior to this comparator. For comparison of efficacy, the primary endpoint is well established and widely accepted by regulatory agencies. However, an important secondary endpoint based on Phase II findings looks very promising. The target dose may have a good opportunity to deliver superiority to the comparator. Furthermore, a lower dose is included in case the target dose may demonstrate potential safety concerns. This Phase III study is designed as a non-inferiority trial with two doses, and two endpoints. This manuscript will illustrate how graphical approach is applied to this design in handling multiple testing issues.

  19. Chemical and mineralogical characterization of the bentonite buffer for the acceptance control procedure in a KBS-3 repository

    Energy Technology Data Exchange (ETDEWEB)

    Karnland, Ola (Clay Technology AB, Lund (Sweden))

    2010-09-15

    The present report concerns the chemical and mineralogical characterization of potential buffer materials. A short mineralogical description of bentonite and montmorillonite is given. The report defines, and exemplifies analyses and tests planned for the acceptance control of the bulk material concerning chemical composition, mineralogical composition, original exchangeable cations, cation exchange capacity (CEC), grain density, specific surface area, granule size, water content. In addition, analyses of the clay fraction, i.e. material with a grain size smaller than 2 mum, are described with respect to chemical composition including layer charge, layer charge distribution and cation exchange capacity (CEC). Results from the report by /Karnland et al. 2006/ concerning the two reference bentonites MX-80 and IBECO RWC are used to illustrate the accuracy and precision of the analyses. For each type of analyses the purpose, technique and, in relevant cases, also limits are discussed briefly. An empirical model for determining swelling pressure is presented and used for the quantification of the expected sealing properties given the limits concerning buffer density and montmorillonite content. For the reference bentonites MX-80, the stipulated montmorillonite content interval from 0.75 to 0.9 gives a pressure interval from 8 to 11 MPa at the nominal saturated density 2,000 kg/m3. The stipulated saturated density interval from 1,950 to 2,050 kg/m3 gives a pressure range from 6 to 15 MPa at the measured montmorillonite content of 83% by weight. The combined effects of the stipulated montmorillonite content interval and saturated density interval lead to a pressure range from 5 to 17 MPa. If the increasing effect of accessory minerals, which is proposed by the model, is not taken into account then the combined pressure range is 3 to 14 MPa

  20. Expose : procedure and results of the joint experiment verification tests

    Science.gov (United States)

    Panitz, C.; Rettberg, P.; Horneck, G.; Rabbow, E.; Baglioni, P.

    The International Space Station will carry the EXPOSE facility accommodated at the universal workplace URM-D located outside the Russian Service Module. The launch will be affected in 2005 and it is planned to stay in space for 1.5 years. The tray like structure will accomodate 2 chemical and 6 biological PI-experiments or experiment systems of the ROSE (Response of Organisms to Space Environment) consortium. EXPOSE will support long-term in situ studies of microbes in artificial meteorites, as well as of microbial communities from special ecological niches, such as endolithic and evaporitic ecosystems. The either vented or sealed experiment pockets will be covered by an optical filter system to control intensity and spectral range of solar UV irradiation. Control of sun exposure will be achieved by the use of individual shutters. To test the compatibility of the different biological systems and their adaptation to the opportunities and constraints of space conditions a profound ground support program has been developed. The procedure and first results of this joint Experiment Verification Tests (EVT) will be presented. The results will be essential for the success of the EXPOSE mission and have been done in parallel with the development and construction of the final hardware design of the facility. The results of the mission will contribute to the understanding of the organic chemistry processes in space, the biological adaptation strategies to extreme conditions, e.g. on early Earth and Mars, and the distribution of life beyond its planet of origin.

  1. Phased Startup Initiative Phase 3 Test Procedure (OCRWM)

    Energy Technology Data Exchange (ETDEWEB)

    PAJUNEN, A.L.

    2000-05-10

    The purpose of this test procedure is to safely operate the Fuel Retrieval System (FRS) and Integrated Water Treatment System (IWTS) with specific fuel canisters, and show that canisters containing fuel can be retrieved from the canister queue, decapped in the Canister Decapper, loaded into the Primary Clean Machine (PCM) for fuel cleaning, fuel sorted on the Process Table, then loaded back into fuel canisters and relocated in Basin Storage. Additional Data are collected during this test, beyond that collected during production operations. These data support qualifying the cleaning performance of the PCM, assessing the quantity of scrap generated during the cleaning, and evaluating the impact of fuel retrieval operations on the Basin water quality. The additional data collected primarily consist of weighing fuel and scrap at selected points in the operation, as well as photographing fuel and scrap as it is processed. The time to perform operations is also monitored for comparison with design predictions. Water quality data are collected to establish a base line to predict the effectiveness of equipment design for control of contamination and visibility during production operation.

  2. Air Traffic Management Technology Demostration: 1 Research and Procedural Testing of Routes

    Science.gov (United States)

    Wilson, Sara R.; Kibler, Jennifer L.; Hubbs, Clay E.; Smail, James W.

    2015-01-01

    NASA's Air Traffic Management Technology Demonstration-1 (ATD-1) will operationally demonstrate the feasibility of efficient arrival operations combining ground-based and airborne NASA technologies. The ATD-1 integrated system consists of the Traffic Management Advisor with Terminal Metering which generates precise time-based schedules to the runway and merge points; Controller Managed Spacing decision support tools which provide controllers with speed advisories and other information needed to meet the schedule; and Flight deck-based Interval Management avionics and procedures which allow flight crews to adjust their speed to achieve precise relative spacing. Initial studies identified air-ground challenges related to the integration of these three scheduling and spacing technologies, and NASA's airborne spacing algorithm was modified to address some of these challenges. The Research and Procedural Testing of Routes human-in-the-loop experiment was then conducted to assess the performance of the new spacing algorithm. The results of this experiment indicate that the algorithm performed as designed, and the pilot participants found the airborne spacing concept, air-ground procedures, and crew interface to be acceptable. However, the researchers concluded that the data revealed issues with the frequency of speed changes and speed reversals.

  3. Factors associated with willingness to accept oral fluid HIV rapid testing among most-at-risk populations in China.

    Directory of Open Access Journals (Sweden)

    Huanmiao Xun

    Full Text Available BACKGROUND: The availability of oral fluid HIV rapid testing provides an approach that may have the potential to expand HIV testing in China, especially among most-a-risk populations. There are few investigations about the acceptability of oral fluid HIV testing among most-at-risk populations in China. METHOD: A cross-sectional study with men who have sex with men (MSM, female sex workers (FSW and voluntary counseling and testing (VCT clients was conducted in three cities of Shandong province, China from 2011 to 2012. Data were collected by face-to-face questionnaire. RESULTS: About 71% of participants were willing to accept the oral fluid HIV rapid testing, and home HIV testing was independently associated with acceptability of the new testing method among MSM, FSW and VCT clients (AOR of 4.46, 3.19 and 5.74, respectively. Independent predictors of oral fluid HIV rapid testing acceptability among MSM were having ever taken an oral fluid HIV rapid test (AOR= 15.25, having ever taken an HIV test (AOR= 2.07, and education level (AOR= 1.74. Engagement in HIV-related risk behaviors (AOR= 1.68 was an independent predictor of acceptability for FSW. Having taken an HIV test (AOR= 2.85 was an independent predictor of acceptability for VCT clients. The primary concern about the oral fluid HIV testing was accuracy. The median price they would pay for the testing ranged from 4.8 to 8.1 U.S. dollars. CONCLUSION: High acceptability of oral fluid HIV rapid testing was shown among most-at-risk populations. Findings provide support for oral rapid HIV testing as another HIV prevention tool, and provide a backdrop for the implementation of HIV home testing in the near future. Appropriate pricing and increased public education through awareness campaigns that address concerns about the accuracy and safety of the oral fluid HIV rapid testing may help increase acceptability and use among most-at-risk populations in China.

  4. 21 CFR 352.72 - General testing procedures.

    Science.gov (United States)

    2010-04-01

    ... results. Excess hair on the back is acceptable if the hair is clipped or shaved. (c) Informed consent... product shall be warmed slightly so that it can be applied volumetrically. On heating, care shall be...

  5. 76 FR 58345 - Energy Conservation Program: Test Procedures for Residential Dishwashers, Dehumidifiers, and...

    Science.gov (United States)

    2011-09-20

    ... descriptor for each covered product, unless the Secretary determines that-- (i) The current test procedures.... 42 FR 39964 (August 8, 1977). Since that time, the dishwasher test procedure has undergone a number... F inlet water. 48 FR 9202 (March 3, 1983). DOE amended the test procedure again in 1984 to...

  6. 10 CFR 431.444 - Test procedures for the measurement of energy efficiency.

    Science.gov (United States)

    2010-01-01

    ... 10 Energy 3 2010-01-01 2010-01-01 false Test procedures for the measurement of energy efficiency. 431.444 Section 431.444 Energy DEPARTMENT OF ENERGY ENERGY CONSERVATION ENERGY EFFICIENCY PROGRAM FOR CERTAIN COMMERCIAL AND INDUSTRIAL EQUIPMENT Small Electric Motors Test Procedures § 431.444 Test procedures for the measurement of energy...

  7. Analysis of the acceptance procedure and quality control a virtual simulation system; Analisis del procedimiento de aceptacion y control de calidad de un sistema de simulacion virtual

    Energy Technology Data Exchange (ETDEWEB)

    Gonzalez Ruiz, C.; Pedrero de Aristizabal, D.; Jimenez Rojas, R.; Garcia Hernandez, M. J.; Ruiz Galan, G.; Ayala Lazaro, R.; Garcia Marcos, R.

    2011-07-01

    Acceptance has been made, determining the reference state, commissioning and implementation of control protocol virtual simulation system consists of an image acquisition unit of computerized tomography (CT), an independent external location laser locator and a simulation module associated with the existing scheduler for clinical dosimetry in radiotherapy. This paper summarizes the path followed in this process, together with the established procedure for periodic monitoring and analysis system of the results obtained in the two years of clinical and control.

  8. Acceptability of rapid oral fluid HIV testing among male injection drug users in Taiwan, 1997 and 2007.

    Science.gov (United States)

    Lyu, Shu-Yu; Morisky, Donald E; Yeh, Ching-Ying; Twu, Shiing-Jer; Peng, Eugene Yu-Chang; Malow, Robert M

    2011-04-01

    Rapid oral fluid HIV testing (rapid oral testing) is in the process of being adapted in Taiwan and elsewhere given its advantages over prior HIV testing methods. To guide this process, we examined the acceptability of rapid oral testing at two time points (i.e., 1997 and 2007) among one of the highest risk populations, male injection drug users (IDUs). For this purpose, an anonymous self-administered survey was completed by HIV-negative IDUs involved in the criminal justice system in 1997 (N (1)=137 parolees) and 2007 (N (2)=106 prisoners). A social marketing model helped guide the design of our questionnaire to assess the acceptability of rapid oral testing. This included assessing a new product, across four marketing dimensions: product, price, promotion, and place. Results revealed that in both 1997 and 2007, over 90% indicated that rapid oral testing would be highly acceptable, particularly if the cost was under US$6, and that a pharmacy would be the most appropriate and accessible venue for selling the rapid oral testing kits. The vast majority of survey respondents believed that the cost of rapid oral testing should be federally subsidized and that television and newspaper advertisements would be the most effective media to advertise for rapid oral testing. Both the 1997 and 2007 surveys suggested that rapid oral HIV testing would be particularly accepted in Taiwan by IDUs after release from the criminal justice system. PMID:21271392

  9. Engineering data transfer test with EDCARS using MIL-R-28002 (Raster). Laboratory Acceptance Test and User Application Test

    Energy Technology Data Exchange (ETDEWEB)

    1992-04-17

    This paper documents the results of a sequence of tests conducted to evaluate the DoD Computer-aided Acquisition and Logistic Support (CALS) data interchange capability of the Air Force Engineering Data Computer-Assisted Retrieval System (EDCARS). The CALS initiative specifies a standard digital interface to streamline the interchange of technical data between the DoD and the commercial sector. The CALS Test Network (CTN) is tasked to conduct tests of military standards which specify this digital interface. The testing results outlined in this report are intended to evaluate the EDCARS systems`s ability to sport CALS data interchanges and establish the level of technical data interoperability implemented at this DoD engineering data repository.

  10. Waste Acceptance Testing of Secondary Waste Forms: Cast Stone, Ceramicrete and DuraLith

    Energy Technology Data Exchange (ETDEWEB)

    Mattigod, Shas V.; Westsik, Joseph H.; Chung, Chul-Woo; Lindberg, Michael J.; Parker, Kent E.

    2011-08-12

    To support the selection of a waste form for the liquid secondary wastes from WTP, Washington River Protection Solutions has initiated secondary-waste-form testing work at Pacific Northwest National Laboratory (PNNL). In anticipation of a down-selection process for a waste form for the Solidification Treatment Unit to be added to the ETF, PNNL is conducting tests on four candidate waste forms to evaluate their ability to meet potential waste acceptance criteria for immobilized secondary wastes that would be placed in the IDF. All three waste forms demonstrated compressive strengths above the minimum 3.45 MPa (500 psi) set as a target for cement-based waste forms. Further, none of the waste forms showed any significant degradation in compressive strength after undergoing thermal cycling (30 cycles in a 10 day period) between -40 C and 60 C or water immersion for 90 days. The three leach test methods are intended to measure the diffusion rates of contaminants from the waste forms. Results are reported in terms of diffusion coefficients and a leachability index (LI) calculated based on the diffusion coefficients. A smaller diffusion coefficient and a larger LI are desired. The NRC, in its Waste Form Technical Position (NRC 1991), provides recommendations and guidance regarding methods to demonstrate waste stability for land disposal of radioactive waste. Included is a recommendation to conduct leach tests using the ANS 16.1 method. The resulting leachability index (LI) should be greater than 6.0. For Hanford secondary wastes, the LI > 6.0 criterion applies to sodium leached from the waste form. For technetium and iodine, higher targets of LI > 9 for Tc and LI > 11 for iodine have been set based on early waste-disposal risk and performance assessment analyses. The results of these three leach tests conducted for a total time between 11days (ASTM C1308) to 90 days (ANS 16.1) showed: (1) Technetium diffusivity: ANSI/ANS 16.1, ASTM C1308, and EPA 1315 tests indicated that

  11. Application of ultrasonic testing as acceptance test for the graphite component of HTTR (High Temperature Test Reactor)

    International Nuclear Information System (INIS)

    Ultrasonic testing method for the nuclear grade graphite was established. The effectiveness of this method was demonstrated through the application for the graphite component of the HTTR. The characteristics of noise echo were also investigated by relating with mechanical properties. (author). 2 refs., 6 figs., 3 tabs

  12. Delinquency and Peer Acceptance in Adolescence: A Within-Person Test of Moffitt's Hypotheses

    Science.gov (United States)

    Rulison, Kelly L.; Kreager, Derek A.; Osgood, D. Wayne

    2014-01-01

    We tested 2 hypotheses derived from Moffitt's (1993) taxonomic theory of antisocial behavior, both of which are central to her explanation for the rise in delinquency during adolescence. We tested whether persistently delinquent individuals become more accepted by their peers during adolescence and whether individuals who abstain from…

  13. Value Acceptance on Adolescent Socialization: A Test of a Cognitive-Functional Theory of Television Effects.

    Science.gov (United States)

    Tan, Alexis; Nelson, Leigh; Dong, Qingwen; Tan, Gerdean

    1997-01-01

    Presents a cognitive-functional explanation of television's influence on Anglo-American, Native American, and Hispanic adolescents' acceptance of values. Finds that adolescents accepted values observed in television when they recognized them (a measure of learning) and when they evaluated the values to be important in "being successful" in the…

  14. AIChe equipment testing procedure centrifugal compressors : a guide to performance evaluation and site testing

    CERN Document Server

    AIChE

    2013-01-01

    With its engineer-tested procedures and thorough explanations, Centrifugal Compressors is an essential text for anyone engaged in implementing new technology in equipment design, identifying process problems, and optimizing equipment performance.  This condensed book presents a step by step approach to preparing for, planning, executing, and analyzing tests of centrifugal compressors, with an emphasis on methods that can be conducted on-site and with an acknowledgement of the strengths and limitations of these methods. The book opens with an extensive and detailed section offering definitions

  15. 9 CFR 147.8 - Procedures for preparing egg yolk samples for diagnostic tests.

    Science.gov (United States)

    2010-01-01

    ... 9 Animals and Animal Products 1 2010-01-01 2010-01-01 false Procedures for preparing egg yolk samples for diagnostic tests. 147.8 Section 147.8 Animals and Animal Products ANIMAL AND PLANT HEALTH... IMPROVEMENT PLAN Blood Testing Procedures § 147.8 Procedures for preparing egg yolk samples for...

  16. Acceptance of referral for partners by clients testing positive for human immunodeficiency virus

    Directory of Open Access Journals (Sweden)

    Netsanet F

    2013-01-01

    Full Text Available Fetene Netsanet,1 Ayalew Dessie21IMA World Health SuddHealth Multi Donor Trust Fund-Basic Package of Health Services Project, Juba, South Sudan; 2United States Agency for International Development, Private Health Sector Program, Abt Associates Inc, Addis Ababa, EthiopiaBackground: Human immunodeficiency virus (HIV-positive individuals who do not disclose their HIV status to their partners are more likely to present late for HIV and acquired immune deficiency syndrome (AIDS care than those who have disclosed their HIV status to their partners. A major area of challenge with regards to HIV counseling for clients is disclosure of their HIV status to their partners. The main methods of partner notification are patient referral, provider referral, contract referral, and outreach assistance. The emphasis on a plausible and comprehensive partner referral strategy for widespread positive case detection in resource-limited countries needs to be thought out and developed.Methods: A qualitative study was conducted among newly HIV-positive clients to identify partners for notification and acceptance of referral by their partners. Health service providers working in HIV testing and counseling clinics were also provided with semistructured questionnaires in order to assess their view towards partner notification strategies for clients testing positive for HIV.Results: Fifteen newly diagnosed HIV-positive clients were counseled to provide referral slips to their partners. All clients agreed and took the referral card. However, only eight were willing and actually provided the card to their partners. Five of the eight partners of clients who tested HIV-positive and who were provided with referral cards responded to the referral and were tested for HIV. Three were positive and two were negative. Nine of 11 counselors did not agree to requesting partner locator information from HIV-positive clients for contractual referral and/or outreach assistance. The findings

  17. 49 CFR 572.166 - Knees and knee impact test procedure.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 7 2010-10-01 2010-10-01 false Knees and knee impact test procedure. 572.166... TRAFFIC SAFETY ADMINISTRATION, DEPARTMENT OF TRANSPORTATION (CONTINUED) ANTHROPOMORPHIC TEST DEVICES Hybrid III Six-Year-Old Weighted Child Test Dummy § 572.166 Knees and knee impact test procedure....

  18. 49 CFR 572.126 - Knees and knee impact test procedure.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 7 2010-10-01 2010-10-01 false Knees and knee impact test procedure. 572.126... TRAFFIC SAFETY ADMINISTRATION, DEPARTMENT OF TRANSPORTATION (CONTINUED) ANTHROPOMORPHIC TEST DEVICES Six-year-old Child Test Dummy, Beta Version § 572.126 Knees and knee impact test procedure. (a)...

  19. 40 CFR 86.150-98 - Refueling test procedure; overview.

    Science.gov (United States)

    2010-07-01

    ... Regulations for 1977 and Later Model Year New Light-Duty Vehicles and New Light-Duty Trucks and New Otto-Cycle... motor vehicle shall be functioning during all procedures in this subpart except: (1) In cases...

  20. Test procedure for the moisture measurement in tyre carcasses

    International Nuclear Information System (INIS)

    The present invention pertains generally to a procedure for the moisture measurement in tyre carcasses with textile insertions by means of neutrons. It allows a permanent control during manufacturing. The invention is applicable to the retreading of tyres

  1. Testing procedural justice theory in Belgium and Sweden

    OpenAIRE

    Van Damme, Anjuli; Pauwels, Lieven

    2013-01-01

    Procedural justice theory assumes that trust in procedural justice and in the effectiveness of the police are important issues for building the legitimacy of the aforementioned institution. Additionally, the perception of police legitimacy, as a result of public trust, is necessary for the recognition of police authority. When citizens recognize the right of the police to determine authority, they are assumed to feel the obligation to obey the police and ultimately comply with the law and coo...

  2. W-026 acceptance test report plant control system software(submittal {number_sign}223.02)

    Energy Technology Data Exchange (ETDEWEB)

    Watson, T.L.

    1997-02-14

    Acceptance Testing of the WRAP 1 Plant Control System software was conducted throughout the construction of WRAP 1 with final testing on the glovebox software being completed in December 1996. The software tests were broken out into five sections; one for each of the four Local Control Units and one for the supervisory software modules. This document contains a completed copy of the software tests along with the applicable test log and completed Exception Test Reports. The ETRs outside the scope of the contractor are not signed off. These will be resolved by the Buyer and all 1280 open issues will be tracked on Buyer`s ETR database pending resolution.

  3. Optimization of the general acceptability though affective tests and response surface methodology of a dry cacao powder mixture based beverage

    Directory of Open Access Journals (Sweden)

    Elena Chau Loo Kung

    2013-09-01

    Full Text Available This research work had as main objective optimizing the general acceptability though affective tests and response surface methodology of a dry cacao powder mixture based beverage. We obtained formulations of mixtures of cacao powder with different concentrations of 15%, 17.5% and 20%, as well as lecithin concentrations of 0.1%; 0.3%; and 0.5% maintaining a constant content of sugar (25 %, Vanillin (1% that included cacao powder with different pH values: natural (pH 5 and alkalinized (pH 6.5 and pH 8 and water by difference to 100%, generating a total of fifteen treatments to be evaluated, according to the Box-Behnen design for three factors. The treatments underwent satisfaction level tests to establish the general acceptability. The treatment that included cacao powder with a concentration of 17.5 %, pH 6.5 and lecithin concentration of 0.3 % obtained the best levels of acceptability. The software Statgraphics Plus 5.1 was used to obtain the treatment with maximum acceptability that corresponded to cacao powder with pH 6.81, with a concentration of 18.24 % and soy lecithin in 0.28% with a tendency to what was obtained in the satisfaction levels tests. Finally we characterized in a physical-chemistry and microbiological way the optimum formulation as well as evaluated sensitively obtaining an acceptability of 6.17.

  4. SFC/SFBMN guidelines update for nuclear cardiology procedures: stress testing in adults and children

    International Nuclear Information System (INIS)

    The guidelines update for nuclear cardiology procedures are studied in this article. We find the minimum technique conditions for the stress testing practice, the recommendations for the different ischemia activation tests, the choice of the stress test. (N.C.)

  5. Acceptability and Barriers to Uptake of Voluntary Counselling and Testing for HIV amongst Pregnant Women in Nigeria

    Directory of Open Access Journals (Sweden)

    Bukola Olateju Omolase

    2010-08-01

    Full Text Available AIM: This study was designed to determine the acceptability of voluntary counselling and testing (VCT for HIV and identify possible barriers to it’s uptake amongst pregnant women in a Nigerian community. METHODS: One hundred and twenty seven consenting pregnant women attending antenatal clinic at Federal Medical Centre, Owo were randomly selected and interviewed by the authors and two trained assistants. The study was conducted between August and October, 2007. The information obtained with the aid of semi –structured questionnaire included their bio data, awareness about VCT and vertical transmission of HIV-AIDS. Other information obtained included acceptability of VCT and barriers to its uptake. RESULT: Majority (85% were aware about vertical transmission of HIV, most (83.5% knew of VCT. They were predominantly willing to accept VCT (77.2% and were ready to recommend VCT to others(76.2%. The main identified barriers to uptake of voluntary counseling were fear of possible outcome (82.7% and risk of divorce (70.9%. CONCLUSION: Most respondents were aware of voluntary counselling and testing for HIV as well as the vertical transmission of HIV. Majority were willing to accept HIV test and recommend same to others. Barriers to uptake of HIV test included fear of possible outcome, lack of felt need, stigmatization and financial constraint. There is need to extend voluntary counselling and testing for HIV to all antenatal patients. [TAF Prev Med Bull 2010; 9(4.000: 309-314

  6. 47 CFR 11.61 - Tests of EAS procedures.

    Science.gov (United States)

    2010-10-01

    ... and 8:30 a.m. They will originate from Local or State Primary sources. The time and script content... monitoring a state or local primary source to participate in testing. Tests should be performed on 10% of all... message when transmitting the required weekly test. (3) Periodic National Tests. National Primary...

  7. Comparison of Real World Energy Consumption to Models and Department of Energy Test Procedures

    Energy Technology Data Exchange (ETDEWEB)

    Goetzler, William [Navigant Consulting, Inc., Burlington, MA (United States); Sutherland, Timothy [Navigant Consulting, Inc., Burlington, MA (United States); Kar, Rahul [Navigant Consulting, Inc., Burlington, MA (United States); Foley, Kevin [Navigant Consulting, Inc., Burlington, MA (United States)

    2011-09-01

    This study investigated the real-world energy performance of appliances and equipment as it compared with models and test procedures. The study looked to determine whether the U.S. Department of Energy and industry test procedures actually replicate real world conditions, whether performance degrades over time, and whether installation patterns and procedures differ from the ideal procedures. The study first identified and prioritized appliances to be evaluated. Then, the study determined whether real world energy consumption differed substantially from predictions and also assessed whether performance degrades over time. Finally, the study recommended test procedure modifications and areas for future research.

  8. Market testing and consumer acceptance of irradiated rice (Oryza sativa indica Linn.)

    International Nuclear Information System (INIS)

    Special grade A fragrant rice (Jasmine rice) of 13% moisture content was obtained from a local miller in Bangkok. Low density polyethylene, 29.5 cm in width x 45 cm in length and 200 micron in thickness, was used to pack the rice with a net weight of 5 kg. The irradiated food label was printed on one side of the bag to comply with food control regulations. The color and the ink for marking were tested for gamma radiation compatibility. A total of 800 bags of rice, with a total gross weight of 4,000 kg, were irradiated at a minimum absorbed dose at 0.5 kGy for insect disinfestation. Radiation treatment was carried out using a multi-purpose, carrier type gamma irradiator (Model JS-8900, Serial No. IR-155) located at the Thai Irradiation Center. Irradiated rice was distributed on a weekly basis to food stores in Bangkok and Pathum Thani, as well as to various governmental organizations and interested individuals. The product was sold at 60 bahts per bag (approx. US$ 2.4) to retailers. Various commercial brands of non-irradiated rice of 5 kg size, were available in the market at 52 to 78 bahts per bag (approx. US $ 2.08 to 3.12), depending on quality and brand name. During the distribution, a leaflet of educational information was given to the consumer. A simple questionnaire used in the marketing trial indicated that 72% of the consumers bought irradiated rice because of the good quality of the product based on visual inspection, and 28% of them were willing to try the new product. Most consumers preferred irradiated rice to chemical treatment (fumigation) for insect disinfestation. However, most consumers were not sure if they would like to buy irradiated rice again unless its cooking quality was acceptable. Market testing of irradiated rice in the upper-class market or supermarket was unsuccessful because of limitations in the sale and service conditions. To meet the requirement of the supermarket retailer, irradiated rice had to be supplied on a monthly basis, with

  9. South African physicians’ acceptance of e-prescribing technology: an empirical test of a modified UTAUT model

    Directory of Open Access Journals (Sweden)

    Jason Cohen

    2013-07-01

    Full Text Available E-prescribing systems hold promise for improving the quality and efficiency of the scripting process. Yet, the use of the technology has been associated with a number of challenges. The diffusion of e-prescribing into physician practices and the consequent realisation of its potential benefits will depend on whether physicians are willing to accept and engage with the technology. This study draws on the Unified Theory of Acceptance and Use of Technology (UTAUT and recent literature on user trust in technology to develop and test a model of the factors influencing South African physicians’ acceptance of e-prescribing. Data was collected from a sample of 72 physicians. Results indicate a general acceptance of e-prescribing amongst physicians who on average reported strong intentions to use e-prescribing technologies if given the opportunity. PLS analysis revealed that physicians’ performance expectancies and perceptions of facilitating conditions had significant direct effects on acceptance whilst trust and effort expectancy had important indirect effects. Social influence and price value perceptions did not add additional explanatory power. The model explained 63% of the variation in physician acceptance.

  10. 77 FR 4203 - Energy Conservation Program: Test Procedures for General Service Fluorescent Lamps, General...

    Science.gov (United States)

    2012-01-27

    ... addressing test procedures for fluorescent and incandescent lamps. 62 FR 29221 (May 29, 1997); 74 FR 31829... previous test procedure rulemaking on incandescent and fluorescent performance testing. 62 FR 29221, 29229... Color Temperature (CCT) of Electric Lamps''). 76 FR 56661, 56662 (September 14, 2011). DOE...

  11. 77 FR 74559 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Water Heaters...

    Science.gov (United States)

    2012-12-17

    ... given water heater. 75 FR 52892, 52895 (August 30, 2010). Specifically, the standby loss testing in the... to consider amendments to DOE's test procedures for residential water heaters. 76 FR 63211 (Oct. 12... Part 430 RIN 1904-AB95 Energy Conservation Program for Consumer Products: Test Procedures...

  12. Empirical Testing of a Theoretical Extension of the Technology Acceptance Model: An Exploratory Study of Educational Wikis

    Science.gov (United States)

    Liu, Xun

    2010-01-01

    This study extended the technology acceptance model and empirically tested the new model with wikis, a new type of educational technology. Based on social cognitive theory and the theory of planned behavior, three new variables, wiki self-efficacy, online posting anxiety, and perceived behavioral control, were added to the original technology…

  13. A Randomized Controlled Trial of Acceptance-Based Behavior Therapy and Cognitive Therapy for Test Anxiety: A Pilot Study

    Science.gov (United States)

    Brown, Lily A.; Forman, Evan M.; Herbert, James D.; Hoffman, Kimberly L.; Yuen, Erica K.; Goetter, Elizabeth M.

    2011-01-01

    Many university students suffer from test anxiety that is severe enough to impair performance. Given mixed efficacy results of previous cognitive-behavior therapy (CBT) trials and a theoretically driven rationale, an acceptance-based behavior therapy (ABBT) approach was compared to traditional CBT (i.e., Beckian cognitive therapy; CT) for the…

  14. Nonparametric tests of collectively rational consumption behavior: an integer programming procedure.

    OpenAIRE

    Cherchye, Laurens; De Rock, Bram; Sabbe, Jeroen; Vermeulen, Frederic

    2008-01-01

    We present an IP-based nonparametric (revealed preference) testing procedure for rational consumption behavior in terms of general collective models, which include consumption externalities and public consumption. An empirical application to data drawn from the Russia Longitudinal Monitoring Survey (RLMS) demonstrates the practical usefulness of the procedure. Finally, we present extensions of the testing procedure to evaluate the goodness of fit of the collective model subject to testing, an...

  15. 14 CFR 61.43 - Practical tests: General procedures.

    Science.gov (United States)

    2010-01-01

    ... practical test: (1) If the aircraft's FAA-approved flight manual requires the pilot flight crew complement be a single pilot, then the applicant must demonstrate single pilot proficiency on the practical test... single pilot, then the applicant must demonstrate single pilot proficiency on the practical test. (3)...

  16. 40 CFR 86.127-12 - Test procedures; overview.

    Science.gov (United States)

    2010-07-01

    ... cycle simulates urban driving behavior with the air conditioner operating. The test consists of engine...) The SC03 air conditioning test is conducted with the air conditioner operating at specified settings and the ambient test conditions of: (i) Air temperature of 95 °F; (ii) 100 grains of water/pound...

  17. 40 CFR 60.46 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ... determined from the average values of Fd and Fc in Method 19 of appendix A of this part, i.e., Foa = 0.209 (Fda/Fca), then the following procedure shall be followed: (A) When Fo is less than 0.97 Foa, then E shall be increased by that proportion under 0.97 Foa, e.g., if Fo is 0.95 Foa, E shall be increased by...

  18. Scoring and Testing Procedures Devoted to Probabilistic Seismic Hazard Assessment

    OpenAIRE

    Albarello, D.; Dipartimento di Scienze Fisiche, della Terra e dell’Ambiente, Università degli Studi di Siena; D’Amico, V.; Istituto Nazionale di Geofisica e Vulcanologia, Sezione Pisa, Pisa, Italia

    2015-01-01

    This review addresses long-term (tens of years) seismic ground-motion forecasting (seismic hazard assessment) in the presence of alternative computational models (the so-called epistemic uncertainty affecting hazard estimates). We review the different approaches that have been proposed to manage epistemic uncertainty in the context of probabilistic seismic hazard assessment (PSHA). Ex-ante procedures (based on the combination of expert judgments about inherent characteristics o...

  19. 49 CFR 219.611 - Test result indicating prohibited alcohol concentration; procedures.

    Science.gov (United States)

    2010-10-01

    ... concentration; procedures. 219.611 Section 219.611 Transportation Other Regulations Relating to Transportation... concentration; procedures. Procedures for administrative handling by the railroad in the event an employee's confirmation test indicates an alcohol concentration of .04 or greater are set forth in § 219.104....

  20. Factors influencing the acceptability of energy policies : A test of VBN theory

    NARCIS (Netherlands)

    Steg, L; Dreijerink, L; Abrahamse, W

    2005-01-01

    This paper examines factors influencing the acceptability of energy policies aimed to reduce the emission of CO2 by households. More specifically, it is Studied to what extent the value-belief-norm theory of environmentalism (VBN theory; Stern, [(2000). Toward a coherent theory of environmentally si

  1. Proof of fault coverage for a formal protocol test procedure

    Science.gov (United States)

    Randall, Michael A.

    1992-12-01

    Due to the speed and complexity of communication networks being designed today, it is imperative to ensure that they operate correctly. Todays fiber optic networks, which can transmit billions of bits per second over thousands of miles, are heavily dependent on sophisticated software and protocols which are becoming increasingly difficult to test. Conformance testing is a method that is used for this purpose: to test the design of a protocol against an implementation of the design. This thesis provides some insight into the conformance testing problem by first providing background on some current protocol test methods and then focusing on a newer method, which is based on a formal protocol specification. A proof is given that demonstrates the method's error detection capabilities. Two well known local area network protocols, Token Bus and Fiber Distributed Data Interface (FDDI), are used as examples to illustrate how the test method is applied to a specification.

  2. A Generalized Benjamini-Hochberg Procedure for Multivariate Hypothesis Testing

    OpenAIRE

    Alishahi, Kasra; Ehyaei, Ahmad Reza; Shojaie, Ali

    2016-01-01

    The introduction of the false discovery rate (FDR) by Benjamini and Hochberg has spurred a great interest in developing methodologies to control the FDR in various settings. The majority of existing approaches, however, address the FDR control for the case where an appropriate univariate test statistic is available. Modern hypothesis testing and data integration applications, on the other hand, routinely involve multivariate test statistics. The goal, in such settings, is to combine the evide...

  3. 40 CFR 61.174 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ... compliance with § 61.172(c), the owner or operator shall conduct emission tests and reduce the test data in... existing copper converter or a copper converter that has an initial startup date preceding the effective... copper converter, initial startup of which occurs after the effective date, or (3) At such other times...

  4. Evaluation of Surface Infiltration Testing Procedures in Permeable Pavement Systems

    Science.gov (United States)

    The ASTM method (ASTM C1701) for measuring infiltration rate of in-place pervious concrete provides limited guidance on how to select testing locations, so research is needed to evaluate how testing sites should be selected and how results should be interpreted to assess surface ...

  5. 40 CFR 86.127-00 - Test procedures; overview.

    Science.gov (United States)

    2010-07-01

    ... for SC03 emissions). The test cycle simulates urban driving behavior with the air conditioner... conditioner operating at specified settings and the ambient test conditions of: (i) Air temperature of 95 °F; (ii) 100 grains of water/pound of dry air (approximately 40 percent relative humidity);...

  6. 49 CFR 325.59 - Measurement procedure; stationary test.

    Science.gov (United States)

    2010-10-01

    ... minutes, to permit the engine radiator fan to automatically disengage when the vehicle's noise emissions... Stationary Test, 40 CFR 202.21. (Table 1 in § 325.7 lists the range of maximum permissible sound level... MOTOR CARRIER NOISE EMISSION STANDARDS Measurement of Noise Emissions; Stationary Test §...

  7. 14 CFR Appendix F to Part 23 - Test Procedure

    Science.gov (United States)

    2010-01-01

    ... materials used in electrical wire and cable insulation and in small parts, materials must be tested either... wire and cable insulation, the wire and cable specimens must be the same size as used in the airplane... specification (make and size) must be tested. The specimen of wire or cable (including insulation) must...

  8. Healing Our Women for Transgender Women: Adaptation, Acceptability, and Pilot Testing.

    Science.gov (United States)

    Collier, Kate L; Colarossi, Lisa G; Hazel, Daphne S; Watson, Kim; Wyatt, Gail E

    2015-10-01

    Healing Our Women (HOW) is a group-level HIV risk-reduction intervention developed to address the role of prior sexual victimization in HIV risk and protective behaviors among HIV-positive women of color. This article describes the process of adapting HOW for transgender women of color in New York City in accordance with CDC guidance for the adaptation of efficacious interventions. Twenty-one transgender women were enrolled in a study to evaluate the acceptability and fidelity of the adapted intervention, and to assess HIV knowledge, depressive symptoms, coping, condom use self-efficacy, and condom use via pre- and post-intervention surveys. We found the adapted program to be feasible to implement and acceptable to participants. We also found significant decreases in depressive symptoms and increases in positive coping from pre- to post-intervention, although replication with a larger sample and a control group comparison is needed to determine efficacy with this population. PMID:26485232

  9. Questionnaire design: carry-over effects of overall acceptance question placement and pre-evaluation instructions on overall acceptance scores in central location tests.

    Science.gov (United States)

    Bastian, Mauresa; Eggett, Dennis L; Jefferies, Laura K

    2015-02-01

    Question placement and usage of pre-evaluation instructions (PEI) in questionnaires for food sensory analysis may bias consumers' scores via carry-over effects. Data from consumer sensory panels previously conducted at a central location, spanning 11 years and covering a broad range of food product categories, were compiled. Overall acceptance (OA) question placement was studied with categories designated as first (the first evaluation question following demographic questions), after nongustation questions (immediately following questions that do not require panelists to taste the product), and later (following all other hedonic and just-about-right [JAR] questions, but occasionally before ranking, open-ended comments, and/or intent to purchase questions). Each panel was categorized as having or not having PEI in the questionnaire; PEI are instructions that appear immediately before the first evaluation question and show panelists all attributes they will evaluate before receiving test samples. Postpanel surveys were administered regarding the self-reported effect of PEI on panelists' evaluation experience. OA scores were analyzed and compared (1) between OA question placement categories and (2) between panels with and without PEI. For most product categories, OA scores tended to be lower when asked later in the questionnaire, suggesting evidence of a carry-over effect. Usage of PEI increased OA scores by 0.10 of a 9-point hedonic scale point, which is not practically significant. Postpanel survey data showed that presence of PEI typically improved the panelists' experience. Using PEI does not appear to introduce a meaningful carry-over effect.

  10. 40 CFR 60.275 - Test methods and procedures.

    Science.gov (United States)

    2010-07-01

    ...: Electric Arc Furnaces Constructed After October 21, 1974, and On or Before August 17, 1983 § 60.275 Test... conducted at least once per day for at least three 6-minute periods when the furnace is operating in...

  11. 77 FR 28673 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Furnace Fans

    Science.gov (United States)

    2012-05-15

    ..., consumer behavior assumptions, and product characteristics (e.g., multi-stage or modulating heating and... Consumer Products: Test Procedures for Residential Furnace Fans; Proposed Rule #0;#0;Federal Register / Vol... 430 RIN 1904-AC21 Energy Conservation Program for Consumer Products: Test Procedures for...

  12. 16 CFR 1610.33 - Test procedures for textile fabrics and film.

    Science.gov (United States)

    2010-01-01

    ... REGULATIONS STANDARD FOR THE FLAMMABILITY OF CLOTHING TEXTILES Rules and Regulations § 1610.33 Test procedures for textile fabrics and film. (a)(1) All textile fabrics (except those with a nitro-cellulose fiber... 16 Commercial Practices 2 2010-01-01 2010-01-01 false Test procedures for textile fabrics and...

  13. 77 FR 31443 - Energy Conservation Program: Test Procedures for Residential Dishwashers, Dehumidifiers, and...

    Science.gov (United States)

    2012-05-25

    ... descriptor for each covered product, unless the Secretary determines that-- (i) the current test procedures... its test procedure for dishwashers in 1977. 42 FR 39964 (August 8, 1977). Since that time, the... and to address dishwashers that use 120 F inlet water. 48 FR 9202 (March 3, 1983). DOE amended...

  14. 40 CFR 1065.15 - Overview of procedures for laboratory and field testing.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 32 2010-07-01 2010-07-01 false Overview of procedures for laboratory and field testing. 1065.15 Section 1065.15 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Applicability and General Provisions §...

  15. 76 FR 65631 - Energy Conservation Program: Test Procedures for Microwave Ovens

    Science.gov (United States)

    2011-10-24

    ... efficiency test procedure for microwave ovens under the Energy Policy and Conservation Act (EPCA). 75 FR... Part 430 RIN 1904-AC26 Energy Conservation Program: Test Procedures for Microwave Ovens AGENCY: Office... microwave-only ovens and combination microwave ovens, including: Food loads representative of consumer...

  16. 78 FR 63823 - Energy Conservation Program: Test Procedures for Television Sets

    Science.gov (United States)

    2013-10-25

    ... consumption of modern TVs. 74 FR 53641. \\1\\ See Energy Conservation Program: Repeal of Test Procedures for Televisions. 74 FR 53640 (Oct. 20, 2009). http://www1.eere.energy.gov/buildings/appliance_standards/pdfs... the proposed television test procedure. 77 FR 2830 (January 19, 2012) (the January 2012 NOPR). In...

  17. 7 CFR 28.603 - Procedures for air flow tests of micronaire reading.

    Science.gov (United States)

    2010-01-01

    ... 7 Agriculture 2 2010-01-01 2010-01-01 false Procedures for air flow tests of micronaire reading... of the United States for Fiber Fineness and Maturity § 28.603 Procedures for air flow tests of...) Air flow instrument complete with accessories to measure the fineness and maturity, in combination,...

  18. 76 FR 647 - Energy Conservation Program: Test Procedures for Electric Motors and Small Electric Motors

    Science.gov (United States)

    2011-01-05

    ... FR 54114. After determining that energy conservation standards for small electric motors would be... of test procedures for certain small electric motors. 71 FR 38807 (July 10, 2006). Pursuant to... adopting test procedures for measuring the energy efficiency of small electric motors. 74 FR 32059....

  19. 10 CFR 431.16 - Test procedures for the measurement of energy efficiency.

    Science.gov (United States)

    2010-01-01

    ... purposes of 10 CFR Part 431 and EPCA, the test procedures for measuring the energy efficiency of an... 10 Energy 3 2010-01-01 2010-01-01 false Test procedures for the measurement of energy efficiency. 431.16 Section 431.16 Energy DEPARTMENT OF ENERGY ENERGY CONSERVATION ENERGY EFFICIENCY PROGRAM...

  20. 46 CFR 197.206 - Substitutes for required equipment, materials, apparatus, arrangements, procedures, or tests.

    Science.gov (United States)

    2010-10-01

    ... particular equipment, material, apparatus, arrangement, procedure, or test is unreasonable or impracticable, the Commandant may permit the use of alternate equipment, material, apparatus, arrangement, procedure, or test to such an extent and upon such condition as will insure, to his satisfaction, a degree...

  1. 76 FR 18428 - Energy Efficiency Program for Certain Commercial and Industrial Equipment: Test Procedures for...

    Science.gov (United States)

    2011-04-04

    .... 71 FR 71340, 71350 (Dec. 8, 2006). ARI Standard 810-2003 references the American National Standards... the clarified energy rate equation published by DOE as part of the previous DOE test procedure (71 FR.... 71 FR 71340, 71350 (Dec. 8, 2006). The DOE test procedure also references ASHRAE Standard 29-1988...

  2. 14 CFR 23.1045 - Cooling test procedures for turbine engine powered airplanes.

    Science.gov (United States)

    2010-01-01

    ... 14 Aeronautics and Space 1 2010-01-01 2010-01-01 false Cooling test procedures for turbine engine powered airplanes. 23.1045 Section 23.1045 Aeronautics and Space FEDERAL AVIATION ADMINISTRATION... CATEGORY AIRPLANES Powerplant Cooling § 23.1045 Cooling test procedures for turbine engine...

  3. 10 CFR 431.193 - Test procedures for measuring energy consumption of distribution transformers.

    Science.gov (United States)

    2010-01-01

    ... distribution transformers. 431.193 Section 431.193 Energy DEPARTMENT OF ENERGY ENERGY CONSERVATION ENERGY EFFICIENCY PROGRAM FOR CERTAIN COMMERCIAL AND INDUSTRIAL EQUIPMENT Distribution Transformers Test Procedures § 431.193 Test procedures for measuring energy consumption of distribution transformers. The...

  4. Building Energy Simulation Test for Existing Homes (BESTEST-EX); Phase 1 Test Procedure: Building Thermal Fabric Cases

    Energy Technology Data Exchange (ETDEWEB)

    Judkoff, R.; Polly, B.; Bianchi, M.; Neymark, J.

    2010-08-01

    The U.S. Department of Energy tasked NREL to develop a process for testing the reliability of models that predict retrofit energy savings, including their associated calibration methods. DOE asked NREL to conduct the work in phases so that a test procedure would be ready should DOE need it to meet legislative requirements related to residential retrofits in FY 2010. This report documents the initial 'Phase 1' test procedure.

  5. Accuracy and Acceptability of Oral Fluid HIV Self-Testing in a General Adult Population in Kenya.

    Science.gov (United States)

    Kurth, Ann E; Cleland, Charles M; Chhun, Nok; Sidle, John E; Were, Edwin; Naanyu, Violet; Emonyi, Wilfred; Macharia, Stephen M; Sang, Edwin; Siika, Abraham M

    2016-04-01

    We evaluated performance, accuracy, and acceptability parameters of unsupervised oral fluid (OF) HIV self-testing (HIVST) in a general population in western Kenya. In a prospective validation design, we enrolled 240 adults to perform rapid OF HIVST and compared results to staff administered OF and rapid fingerstick tests. All reactive, discrepant, and a proportion of negative results were confirmed with lab ELISA. Twenty participants were video-recorded conducting self-testing. All participants completed a staff administered survey before and after HIVST to assess attitudes towards OF HIVST acceptability. HIV prevalence was 14.6 %. Thirty-six of the 239 HIVSTs were invalid (15.1 %; 95 % CI 11.1-20.1 %), with males twice as likely to have invalid results as females. HIVST sensitivity was 89.7 % (95 % CI 73-98 %) and specificity was 98 % (95 % CI 89-99 %). Although sensitivity was somewhat lower than expected, there is clear interest in, and high acceptability (94 %) of OF HIV self-testing. PMID:26438487

  6. Soil washing physical separations test procedure - 300-FF-1 operable unit

    International Nuclear Information System (INIS)

    This procedure provides the operations approach, a field sampling plan, and laboratory procedures for a soil washing test to be conducted by Alternative Remedial Technologies, Inc. (ART) in the 300-FF-1 area at the Hanford site. The open-quotes Quality Assurance Project Plan for the Soil Washing Physical Separations Test, 300-FF-1 Operable Unit,close quotes Hanford, Washington, Alternative Remedial Technologies, Inc., February 1994 (QAPP) is provided in a separate document that presents the procedural and organizational guidelines for this test. This document describes specifications, responsibilities, and general procedures to be followed to conduct physical separation soil treatability tests in the North Process Pond of the 300-FF-1 Operable Unit (OU) at the Hanford Site. These procedures are based on the open-quotes 300-FF-1 Physical Separations CERCLA Treatability Test Plan, DOE/RL 92-2l,close quotes (DOE-RL 1993)

  7. Soil washing physical separations test procedure - 300-FF-1 operable unit

    Energy Technology Data Exchange (ETDEWEB)

    Belden, R.D.

    1993-10-08

    This procedure provides the operations approach, a field sampling plan, and laboratory procedures for a soil washing test to be conducted by Alternative Remedial Technologies, Inc. (ART) in the 300-FF-1 area at the Hanford site. The {open_quotes}Quality Assurance Project Plan for the Soil Washing Physical Separations Test, 300-FF-1 Operable Unit,{close_quotes} Hanford, Washington, Alternative Remedial Technologies, Inc., February 1994 (QAPP) is provided in a separate document that presents the procedural and organizational guidelines for this test. This document describes specifications, responsibilities, and general procedures to be followed to conduct physical separation soil treatability tests in the North Process Pond of the 300-FF-1 Operable Unit (OU) at the Hanford Site. These procedures are based on the {open_quotes}300-FF-1 Physical Separations CERCLA Treatability Test Plan, DOE/RL 92-2l,{close_quotes} (DOE-RL 1993).

  8. Single specimen fracture toughness determination procedure using instrumented impact test

    International Nuclear Information System (INIS)

    In the study a new single specimen test method and testing facility for evaluating dynamic fracture toughness has been developed. The method is based on the application of a new pendulum type instrumented impact tester equipped with and optical crack mouth opening displacement (COD) extensometer. The fracture toughness measurement technique uses the Double Displacement Ratio (DDR) method, which is based on the assumption that the specimen is deformed as two rigid arms that rotate around an apparent centre of rotation. This apparent moves as the crack grows, and the ratio of COD versus specimen displacement changes. As a consequence the onset ductile crack initiation can be detected on the load-displacement curve. Thus, an energy-based fracture toughness can be calculated. In addition the testing apparatus can use specimens with the Double ligament size as compared with the standard Charpy specimen which makes the impact testing more appropriate from the fracture mechanics point of view. The novel features of the testing facility and the feasibility of the new DDR method has been verified by performing an extensive experimental and analytical study. (99 refs., 91 figs., 27 tabs.)

  9. Operability test procedure for 244-U DCRT. Revision 1

    Energy Technology Data Exchange (ETDEWEB)

    Erhart, M.F.

    1995-03-13

    This OPT will insure the operability of various systems and their general equipment, used in the operation of the 244-U DCRT. Systems that will be tested include the following: Leak Detection Systems; Heat Trace System; Vessel Temperature Measurement System; Dip Tube Water System; Weight Factor and Specific Gravity System; Instrument Air System; Vessel Liquid Level System; Vessel Liquid Transfer System; and Exhauster Differential Pressure Indicating Controller System. The tests will cover checks on equipment operation and electrical interlocks, but will not cover checks on associated annunciator alarms which are covered in a separate OTP.

  10. International Space Station Environmental Control and Life Support System Acceptance Testing for Node 1 Temperature and Humidity Control Subsystem

    Science.gov (United States)

    Williams, David E.

    2011-01-01

    The International Space Station (ISS) Node 1 Environmental Control and Life Support (ECLS) System is comprised of five subsystems: Atmosphere Control and Storage (ACS), Atmosphere Revitalization (AR), Fire Detection and Suppression (FDS), Temperature and Humidity Control (THC), and Water Recovery and Management (WRM). This paper will provide a summary of the Node 1 ECLS THC subsystem design and a detailed discussion of the ISS ECLS Acceptance Testing methodology utilized for this subsystem.The International Space Station (ISS) Node 1 Environmental Control and Life Support (ECLS) System is comprised of five subsystems: Atmosphere Control and Storage (ACS), Atmosphere Revitalization (AR), Fire Detection and Suppression (FDS), Temperature and Humidity Control (THC), and Water Recovery and Management (WRM). This paper will provide a summary of the Node 1 ECLS THC subsystem design and a detailed discussion of the ISS ECLS Acceptance Testing methodology utilized for this subsystem.

  11. Acceptability and Feasibility of Sexually Transmitted Infection Testing and Treatment among Pregnant Women in Gaborone, Botswana, 2015

    Directory of Open Access Journals (Sweden)

    Adriane Wynn

    2016-01-01

    Full Text Available Introduction. Chlamydia trachomatis (CT, Neisseria gonorrhoeae (NG, and Trichomonas vaginalis (TV are curable sexually transmitted infections (STIs that can cause adverse maternal and birth outcomes. Most countries do not conduct routine testing during antenatal care. We present data on the acceptability and feasibility of testing and treating pregnant women for STIs in an antenatal clinic in Gaborone, Botswana. Materials and Methods. We offered CT, NG, and TV testing using self-collected vaginal swabs to eligible pregnant women. Participants received same-day test results. Those who tested positive were given treatment. Results. Among the 225 women who were eligible and recruited, 200 (89% agreed to participate. The median age of our study sample was 30 years; most were unmarried (77%, with a median gestational age of 27 weeks and a 23% HIV prevalence. All participants received their results with at least 72% (n=143 on the same day. Thirty participants (15% tested positive for an STI, all were treated, and 24 (80% were treated on the same day. Conclusion. The acceptability of STI testing was high, and the intervention was feasible. This study provides support for continued research into STI prevalence, cost-effectiveness, and the association of STIs with adverse maternal and infant outcomes.

  12. 16 CFR 1632.4 - Mattress test procedure.

    Science.gov (United States)

    2010-01-01

    ... room. (ii) Tape edge. Each burning cigarette shall be placed in the depression between the mattress top... mattress tests. If there is only a seam or no depression at the edge, support the cigarettes in place along... directly over the thread or in the depression created by the quilting process on the half of the...

  13. New testing procedures of a capacitive deionization reactor.

    Science.gov (United States)

    García-Quismondo, Enrique; Gómez, Roberto; Vaquero, Fernando; Cudero, Ana López; Palma, Jesús; Anderson, Marc

    2013-05-28

    Currently, according to conventional charge-discharge profiles, energy consumed in charging Capacitive Deionization (CDI) systems is always a function of different parameters (current used for charging, capacitance and current employed for discharging) making it difficult to separate the effect of these parameters on CDI performance and energy efficiency. Thus, energy efficiencies are strongly influenced by the current in the preceding charge or discharge stage of the process. We find consistently that this phenomenon, which to our knowledge has not been addressed in previous CDI communications, is much more intense when different currents are applied for each of the charging and discharging cycles. The investigation reported here provides a mechanistic analysis of the operational aspects of CDI and develops a new procedure that allows for a precise evaluation of performance and energy efficiency. Furthermore, the model developed here allows one to separate charge and discharge cycles, and therefore contributes to the possibility of defining an operational mode for real-world devices in which effective separation of deionization and regeneration steps needs to be implemented. This method of analysis could be useful not only for CDI but also for other electrochemical systems such as in secondary batteries and supercapacitors where charge and discharge are typically employed. PMID:23591701

  14. A Proposed Standardized Testing Procedure for Autonomous Ground Vehicles

    OpenAIRE

    Alberi, Thomas James

    2008-01-01

    Development of unmanned vehicles will increase as the need to save lives rises. In both military and civilian applications, humans can be taken out of the loop through the implementation of safe and intelligent autonomous vehicles. Although hardware and software development continue to play a large role in the autonomous vehicle industry, validation of these systems will always be necessary. The ability to test these vehicles thoroughly and efficiently will ensure their proper and flawless...

  15. Development of testing procedure for welded steel tubes

    OpenAIRE

    Hameed, Saqib

    2010-01-01

    The objective of this study is to perform physical tests on T-joint welded tubes to determine the axial static load behavior and compared them with the analytical and numerical calculations. Three experiments are performed and it is examined that typical collapse modes develop in the section of tube and associated Von-Mises stresses (distortion energy) are compared with the analytical method and numerical simulations. The comparisons with the experimental results are satisfactory. The validat...

  16. 49 CFR 572.136 - Knees and knee impact test procedure.

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 7 2010-10-01 2010-10-01 false Knees and knee impact test procedure. 572.136... TRAFFIC SAFETY ADMINISTRATION, DEPARTMENT OF TRANSPORTATION (CONTINUED) ANTHROPOMORPHIC TEST DEVICES Hybrid III 5th Percentile Female Test Dummy, Alpha Version § 572.136 Knees and knee impact test...

  17. Comparison of RFFIT tests with different standard sera and testing procedures.

    Science.gov (United States)

    Yu, Peng-cheng; Noguchi, Akira; Inoue, Satoshi; Tang, Qing; Rayner, Simon; Liang, Guo-dong

    2012-06-01

    The World Health Organization (WHO) standard assay for determining antibody level is the rapid fluorescent focus inhibition test (RFFIT) and is used to determine the degree of immunity after vaccination against rabies. To compare the difference in RFFIT results between the laboratories of The National Institute of Infectious Disease in Japan (NIID) and the Chinese Centre for Disease Control (CCDC) as well the influence of the choice of standard serum (STD) for the detection, the two laboratories detection methods were simultaneously manipulated by RFFIT. The reference serums used in NIID and the WHO standard serum used in CCDC were compared in the same RFFIT detection to determine the titer of four sera samples C1, S1, S2 and S4 in parallel, and the titers of the detected sera samples were calculated using the standard formula for neutralizing antibody titer. No significant difference was found in RFFIT methods from the two laboratories and the RFFIT testing procedures of the two laboratories have good consistency. However, different titers were obtained with the tentative internal standard serum (TI-STD) produced by adjusting to 2.0 IU of WHO standard serum in NIID and the WHO STD. The titer determined with the TI-STD was higher than that determined with WHO STD, This difference appears to be significant and requires further investigation.

  18. Comparison of RFFIT Tests with Different Standard Sera and Testing Procedures

    Institute of Scientific and Technical Information of China (English)

    Peng-cheng Yu; Akira Noguchi; Satoshi Inoue; Qing Tang; Simon Rayner; Guo-dong Liang

    2012-01-01

    The World Health Organization (WHO) standard assay for determining antibody level is the rapid fluorescent focus inhibition test (RFFIT) and is used to determine the degree of immunity after vaccination against rabies.To compare the difference in RFFIT results between the laboratories of The National Institute of Infectious Disease in Japan (NIID) and the Chinese Centre for Disease Control (CCDC) as well the influence of the choice of standard serum (STD) for the detection,the two laboratories detection methods were simultaneously manipulated by RFFIT.The reference serums used in NIID and the WHO standard serum used in CCDC were compared in the same RFFIT detection to determine the titer of four sera samples C1,S1,S2 and S4 in parallel,and the titers of the detected sera samples were calculated using the standard formula for neutralizing antibody titer.No significant difference was found in RFFIT methods from the two laboratories and the RFFIT testing procedures of the two laboratories have good consistency.However,different titers were obtained with the tentative internal standard serum (TI-STD) produced by adjusting to 2.0 IU of WHO standard serum in NIID and the WHO STD.The titer determined with the TI-STD was higher than that determined with WHO STD,This difference appears to be significant and requires further investigation.

  19. Comparison of RFFIT tests with different standard sera and testing procedures.

    Science.gov (United States)

    Yu, Peng-cheng; Noguchi, Akira; Inoue, Satoshi; Tang, Qing; Rayner, Simon; Liang, Guo-dong

    2012-06-01

    The World Health Organization (WHO) standard assay for determining antibody level is the rapid fluorescent focus inhibition test (RFFIT) and is used to determine the degree of immunity after vaccination against rabies. To compare the difference in RFFIT results between the laboratories of The National Institute of Infectious Disease in Japan (NIID) and the Chinese Centre for Disease Control (CCDC) as well the influence of the choice of standard serum (STD) for the detection, the two laboratories detection methods were simultaneously manipulated by RFFIT. The reference serums used in NIID and the WHO standard serum used in CCDC were compared in the same RFFIT detection to determine the titer of four sera samples C1, S1, S2 and S4 in parallel, and the titers of the detected sera samples were calculated using the standard formula for neutralizing antibody titer. No significant difference was found in RFFIT methods from the two laboratories and the RFFIT testing procedures of the two laboratories have good consistency. However, different titers were obtained with the tentative internal standard serum (TI-STD) produced by adjusting to 2.0 IU of WHO standard serum in NIID and the WHO STD. The titer determined with the TI-STD was higher than that determined with WHO STD, This difference appears to be significant and requires further investigation. PMID:22684473

  20. Questionnaire design: carry-over effects of overall acceptance question placement and pre-evaluation instructions on overall acceptance scores in central location tests.

    Science.gov (United States)

    Bastian, Mauresa; Eggett, Dennis L; Jefferies, Laura K

    2015-02-01

    Question placement and usage of pre-evaluation instructions (PEI) in questionnaires for food sensory analysis may bias consumers' scores via carry-over effects. Data from consumer sensory panels previously conducted at a central location, spanning 11 years and covering a broad range of food product categories, were compiled. Overall acceptance (OA) question placement was studied with categories designated as first (the first evaluation question following demographic questions), after nongustation questions (immediately following questions that do not require panelists to taste the product), and later (following all other hedonic and just-about-right [JAR] questions, but occasionally before ranking, open-ended comments, and/or intent to purchase questions). Each panel was categorized as having or not having PEI in the questionnaire; PEI are instructions that appear immediately before the first evaluation question and show panelists all attributes they will evaluate before receiving test samples. Postpanel surveys were administered regarding the self-reported effect of PEI on panelists' evaluation experience. OA scores were analyzed and compared (1) between OA question placement categories and (2) between panels with and without PEI. For most product categories, OA scores tended to be lower when asked later in the questionnaire, suggesting evidence of a carry-over effect. Usage of PEI increased OA scores by 0.10 of a 9-point hedonic scale point, which is not practically significant. Postpanel survey data showed that presence of PEI typically improved the panelists' experience. Using PEI does not appear to introduce a meaningful carry-over effect. PMID:25604650

  1. Test procedures for polyester immobilized salt-containing surrogate mixed wastes

    Energy Technology Data Exchange (ETDEWEB)

    Biyani, R.K.; Hendrickson, D.W.

    1997-07-18

    These test procedures are written to meet the procedural needs of the Test Plan for immobilization of salt containing surrogate mixed waste using polymer resins, HNF-SD-RE-TP-026 and to ensure adequacy of conduct and collection of samples and data. This testing will demonstrate the use of four different polyester vinyl ester resins in the solidification of surrogate liquid and dry wastes, similar to some mixed wastes generated by DOE operations.

  2. Analytical Chemistry Laboratory (ACL) procedure compendium. Volume 6, Physical testing

    Energy Technology Data Exchange (ETDEWEB)

    1993-08-01

    This volume contains the interim change notice for physical testing. Covered are: properties of solutions, slurries, and sludges; rheological measurement with cone/plate viscometer; % solids determination; particle size distribution by laser scanning; penetration resistance of radioactive waste; operation of differential scanning calorimeter, thermogravimetric analyzer, and high temperature DTA and DSC; sodium rod for sodium bonded fuel; filling SP-100 fuel capsules; sodium filling of BEATRIX-II type capsules; removal of alkali metals with ammonia; specific gravity of highly radioactive solutions; bulk density of radioactive granular solids; purification of Li by hot gettering/filtration; and Li filling of MOTA capsules.

  3. Kilowatt Isotope Power System: test procedure for the ground demonstration system. 77-KIPS-63

    Energy Technology Data Exchange (ETDEWEB)

    1977-12-02

    This test procedure provides a detailed description of the verification methods which shall be used during development, performance and endurance testing to be conducted on the Kilowatt Isotope Power System (KIPS) Development Ground Demonstration Systems, P/N's 722487 and 723829 and Ground Demonstration System (GDS), P/N 725100. The objectives of this testing are to demonstrate system performance in a controlled environment; to verify results of performance predictions; to verify results compiled by component testing performed per Sundstrand Test Procedures and Test Specifications; and to isolate full scale operational characteristics for evaluation.

  4. A urinary test procedure for identification of cannabidiol in patients undergoing medical therapy with marijuana

    Directory of Open Access Journals (Sweden)

    Wertlake PT

    2016-02-01

    Full Text Available Paul T Wertlake, Michael D Henson Pacific Toxicology Laboratories, Chatsworth, CA, USA Abstract: Marijuana is classified by the Drug Enforcement Agency (DEA as Schedule I, drugs having no accepted medical value. Twenty-three states and the District of Columbia have legalized medical marijuana. This conflict inhibits physicians from prescribing marijuana and the systematic study of marijuana in medical care. This study concerns the use of the clinical laboratory as a resource for physicians recommending cannabidiol (CBD to patients, or for patients using medical marijuana. Marijuana containing delta-9-tetrahydrocannabinol (THC is psychoactive. CBD is not psychoactive. CBD is reported to have medical benefit for seizure control, neurologic disorders including multiple sclerosis, neuropathic pain and pain associated with cancer. Use of opiates leads to increasing dosage over time that may cause respiratory depression. The Medical Board of California has termed this a serious public health crisis of addiction, overdose, and death. Is it feasible that CBD might alleviate persistent, severe pain and therefore diminished opiate use? Further study is needed to determine medical effectiveness of CBD including the effect on concurrent opiate therapy due to competition for receptor sites. This study is the application of a gas chromatography mass spectrometry procedure adapted for use in our laboratory, to detect CBD in urine. The intended use is as a tool for physicians to assess that marijuana being used by a patient is of a composition likely to be medically effective. A law ensuring physicians freedom from federal prosecution would provide confidence essential to formal study of medical uses of marijuana and treatment of clinical problems. Detection of CBD in a urine sample would be a convenient test for such confirmation. Keywords: laboratory test, assay, medical management 

  5. Test procedures and instructions for Hanford complexant concentrate supernatant cesium removal using CST

    Energy Technology Data Exchange (ETDEWEB)

    Hendrickson, D.W.

    1997-01-08

    This document provides specific test procedures and instructions to implement the test plan for the preparation and conduct of a cesium removal test, using Hanford Complexant Concentrate supernatant liquor from tank 241-AN-107, in a bench-scale column. The cesium sorbent to be tested is crystalline silicotitanate. The test plan for which this provides instructions is WHC-SD-RE-TP-023, Hanford Complexant Concentrate Supernatant Cesium Removal Test Plan.

  6. Test procedures and instructions for Hanford tank waste supernatant cesium removal

    Energy Technology Data Exchange (ETDEWEB)

    Hendrickson, D.W., Westinghouse Hanford

    1996-05-31

    This document provides specific test procedures and instructions to implement the test plan for the preparation and conduct of a cesium removal test using Hanford Double-Shell Slurry Feed supernatant liquor from tank 251-AW-101 in a bench-scale column.Cesium sorbents to be tested include resorcinol-formaldehyde resin and crystalline silicotitanate. The test plan for which this provides instructions is WHC-SD-RE-TP-022, Hanford Tank Waste Supernatant Cesium Removal Test Plan.

  7. Determinants of disulfiram acceptance among alcohol patients: a test of the theory of reasoned action.

    Science.gov (United States)

    Brubaker, R G; Prue, D M; Rychtarik, R G

    1987-01-01

    The utility of disulfiram in the treatment of alcohol abuse has been limited by the high frequency with which clients refuse the regimen. The present study identified variables influencing disulfiram acceptance/rejection within the framework of the theory of reasoned action. Eighty male inpatients on an alcohol treatment unit completed a questionnaire operationalizing the components of the model. The theoretical components were then used to predict requests for the drug. Multiple regression analyses revealed significant correlations among components of the model consistent with the assumptions of the theory. In addition, significant differences were found between disulfiram requestors and rejectors in terms of behavioral and normative beliefs about the consequences of being placed on the drug. Implications of these results for the design of effective interventions are discussed.

  8. Acceptability of HIV/AIDS Counseling and Testing among Premarital Couples in China

    Science.gov (United States)

    Wu, Zunyou; Rou, Keming; Xu, Chen; Lou, Wei; Detels, Roger

    2005-01-01

    Premarital counseling is required for couples wishing to be married in China. The counseling primarily provides information about contraception. We evaluated adding premarital HIV/AIDS counseling and voluntary HIV testing to the standard counseling. The test was offered free to one group and at the standard cost to the other. The proportion of…

  9. Test Procedure for Axially Loaded Bucket Foundations in Sand (Large Yellow Box)

    DEFF Research Database (Denmark)

    Vaitkunaite, Evelina

    This is a practical guide for preparing the soil, running a CPT test, installing a scaled bucket foundation model and running a test in the large yellow sand box cos(Kristina) in the geotechnical laboratory at Aalborg University. The test procedure is used for the examination of statically...... and cyclically axially loaded bucket foundation model In dense sand. The foundation model in scale of approximately 1:10 compared to the prototype size. The guide describes the step-by-step procedure for tests with and without surface pressure. A detailed description of test setup using the large yellow sand box...

  10. 40 CFR 63.772 - Test methods, compliance procedures, and compliance demonstrations.

    Science.gov (United States)

    2010-07-01

    ... Method for Determination of Gaseous Organic Compounds by Direct Interface Gas Chromatography-Mass... its use by the procedures specified in Method 21, 40 CFR part 60, appendix A. (4) Calibration gases... Oil and Natural Gas Production Facilities § 63.772 Test methods, compliance procedures, and...

  11. 76 FR 31749 - Energy Conservation Program for Certain Consumer Appliances: Test Procedures for Battery Chargers...

    Science.gov (United States)

    2011-06-01

    ... procedure final rule (74 FR 13318 (March 27, 2009)), did not measure battery charger energy consumption in... time. 71 FR 71340, 71360. The battery charger active mode test procedure in today's final rule removes... definitions into appendices Y and Z. 74 FR 13318. In today's final rule, DOE amends the battery charger...

  12. 75 FR 57556 - Energy Conservation Program for Consumer Products: Test Procedure for Residential Clothes Washers

    Science.gov (United States)

    2010-09-21

    ... technologies not covered by the current procedure; (2) more accurately reflect current consumer behavior and... amendments are based on recent data that more accurately describe current consumer behavior and updated... Energy 10 CFR Part 430 Energy Conservation Program for Consumer Products: Test Procedure for...

  13. 75 FR 57555 - Energy Conservation Program for Consumer Products: Test Procedure for Residential Clothes Washers

    Science.gov (United States)

    2010-09-21

    ... technologies not covered by the current procedure; (2) more accurately reflect current consumer behavior and... amendments are based on recent data that more accurately describe current consumer behavior and updated... Energy 10 CFR Part 430 Energy Conservation Program for Consumer Products: Test Procedure for...

  14. 16 CFR 1611.33 - Test procedures for textile fabrics and film.

    Science.gov (United States)

    2010-01-01

    ... Clothing Textiles, Commercial Standard 191-53”. (2) Notwithstanding the provisions of paragraph (a)(1) of... 16 Commercial Practices 2 2010-01-01 2010-01-01 false Test procedures for textile fabrics and film... procedures for textile fabrics and film. (a)(1) All textile fabrics (except those with a...

  15. 75 FR 64621 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Furnaces and...

    Science.gov (United States)

    2010-10-20

    ... descriptor for each covered product, unless the Secretary determines that-- (i) The current test procedures... rule published in the Federal Register on May 12, 1997. 62 FR 26140. This procedure establishes a means...). 74 FR 36959. DOE's proposal was presented and explained at a public meeting on August 18, 2009 at...

  16. 77 FR 2829 - Energy Conservation Program: Test Procedure for Television Sets

    Science.gov (United States)

    2012-01-19

    ..., Federal Communications Commission (FCC) transition from analog to digital broadcast transmissions for TVs... Electrochemical Commission's (IEC's) test procedure IEC Standard 62087- 2008, ``Methods of measurement for the... procedure for TVs on June 29, 1979, codified at 10 CFR part 430, subpart B, appendix H. 44 FR 37938. In...

  17. 75 FR 2122 - Additional Guidance Regarding Application of Current Procedures for Testing Energy Consumption of...

    Science.gov (United States)

    2010-01-14

    ...) standard HRF-1-1979 (HRF-1). See Appendix A1, section 2.2; 47 FR 34517 (Aug. 10, 1982). HRF-1 specifies... Additional Guidance Regarding Application of Current Procedures for Testing Energy Consumption of... procedures to measure the energy consumption of certain consumer products. See 42 U.S.C. 6293, 6295(r)....

  18. 76 FR 63211 - Energy Efficiency Program: Test Procedures for Residential Water Heaters, Direct Heating...

    Science.gov (United States)

    2011-10-12

    ... procedure requirements under EPCA.\\3\\ 75 FR 52892. DOE published a supplemental notice of proposed... rounding and sampling. 76 FR 56347. However, that rulemaking was limited to the proposed test procedure... chapter 7 of the technical support document (TSD) for DOE's April 16, 2010 final rule (75 FR 20112)...

  19. Acceptability of HIV/AIDS testing among pre-marital couples in Iran (2012

    Directory of Open Access Journals (Sweden)

    Jamshid Ayatollahi

    2014-01-01

    Full Text Available Background: Human immunodeficiency virus (HIV/acquired immune deficiency syndrome (AIDS is a lifestyle-related disease. This disease is transmitted through unprotected sex, contaminated needles, infected blood transfusion and from mother to child during pregnancy and delivery. Prevention of infection with HIV, mainly through safe sex and needle exchange programmes is a solution to prevent the spread of the disease. Knowledge about HIV state helps to prevent and subsequently reduce the harm to the later generation. The purpose of this study was to assess the willingness rate of couples referred to the family regulation pre-marital counselling centre for performing HIV test before marriage in Yazd. Patients and Methods: In this descriptive study, a simple random sampling was done among people referred to Akbari clinic. The couples were 1000 men and 1000 women referred to the premarital counselling centre for pre-marital HIV testing in Yazd in the year 2012. They were in situations of pregnancy, delivery or nursing and milking. The data were analyzed using Statistical Package for the Social Sciences (SPSS software and chi-square statistical test. Results: There was a significant statistical difference between the age groups about willingness for HIV testing before marriage (P < 0.001 and also positive comments about HIV testing in asymptomatic individuals (P < 0.001. This study also proved a significant statistical difference between the two gender groups about willingness to marry after HIV positive test of their wives. Conclusion: The willingness rate of couples to undergo HIV testing before marriage was significant. Therefore, HIV testing before marriage as a routine test was suggested.

  20. Is the nacelle mounted anemometer an acceptable option in performance testing?

    Energy Technology Data Exchange (ETDEWEB)

    Dahlberg, J.A. [FFA, The Aeronautical Research Inst. of Sweden, Bromma (Sweden); Frandsen, S.; Madsen, H.A; Antoniou, I.; Friis Pedersen, T. [Risoe National Lab., Roskilde (Denmark); Hunter, R. [RES, Renewable Energy Systems, Glasgow, Scotland (United Kingdom); Klug, H. [DEWI, Wilhelmshaven (Germany)

    1999-03-01

    Although the nacelle anemometer method has been used for power verification purposes for several years, it is only relatively recently that a full understanding of its limitations has emerged. the technique is totally dependent upon the assumption that the nacelle to free wind speed relationship established for a reference turbine in free air can be applied universally to other turbines. Facts emerged from research projects have shown that this assumption is unjustified. In the present paper facts are presented of which some have not been identified nor presented before. E.g. the effect related to wake conditions is novel as a phenomena and the size of the effect can be considerable. The analysis shows that the total error caused by the effects considered in this paper can, in the worst case reach unacceptable high values, 24%, but by taking precautionary measures the errors can be kept at acceptable low levels, 4%. It is found probable that the future use of nacelle anemometry for power performance verification will be subject to strong restrictions. (au)

  1. 9 CFR 77.33 - Testing procedures for tuberculosis in captive cervids.

    Science.gov (United States)

    2010-01-01

    ... 9 Animals and Animal Products 1 2010-01-01 2010-01-01 false Testing procedures for tuberculosis in captive cervids. 77.33 Section 77.33 Animals and Animal Products ANIMAL AND PLANT HEALTH INSPECTION... tests, the testing veterinarian must submit a report to cooperating State and Federal animal...

  2. 75 FR 16957 - Energy Conservation Program: Test Procedures for Battery Chargers and External Power Supplies

    Science.gov (United States)

    2010-04-02

    ... Framework Document for Battery Chargers and External Power Supplies on June 4, 2009. 74 FR 26816. DOE then... Test Procedure B. Review of Battery Charger and External Power Supply Standby Mode and Off Mode Test...) Battery Preparation (e) Reversed Testing Order (f) End of Discharge for Other Chemistries C. Review...

  3. Introduction procedure of marketing testing innovative products in practical activity of industrial enterprises

    Directory of Open Access Journals (Sweden)

    E.I. Nagornyy

    2010-12-01

    Full Text Available The article considers the relevance of implementing marketing testing procedures in practical activities of industrial enterprises that develop, manufacture and sales of new or innovative products.

  4. 75 FR 54048 - Energy Efficiency Program: Test Procedure for Televisions; Request for Information and Request...

    Science.gov (United States)

    2010-09-03

    .... Introduction DOE adopted a test procedure for televisions (TVs) on June 29, 1979, as described in 44 FR 37938... televisions (74 FR 53640). As of June 12, 2009, the ``Digital Transition and Public Safety Act of...

  5. Animal testing, 3R models and regulatory acceptance : Technology transition in a risk-averse context

    OpenAIRE

    Schiffelers, M.J.W.A.

    2016-01-01

    Risk avoidance has resulted in a broad range of regulations to guarantee the safety of products such as pharmaceuticals and chemicals. Many of these regulations rely on animal tests. About 3 million laboratory animals are used annually in Europe to meet such regulatory requirements.Regulatory animal testing raises concerns for scientific, ethical and economic reasons. There are many 3R initiatives to replace, reduce, or refine laboratory animal use. A broad range of 3R models is available and...

  6. HIV coping self-efficacy: a key to understanding stigma and HIV test acceptance among incarcerated men in Jamaica.

    Science.gov (United States)

    Andrinopoulos, Katherine; Kerrigan, Deanna; Figueroa, J Peter; Reese, Richard; Ellen, Jonathan M

    2010-03-01

    Although prisons have been noted as important venues for HIV testing, few studies have explored the factors within this context that may influence HIV test acceptance. Moreover, there is a dearth of research related to HIV and incarcerated populations in middle and low-income countries, where both the burden of HIV and the number of people incarcerated is higher compared to high-income countries. This study explores the relationship between HIV coping self-efficacy, HIV-related stigma, and HIV test acceptance in the largest prisons in Jamaica. A random sample of inmates (n=298) recruited from an HIV testing demonstration project were asked to complete a cross-sectional quantitative survey. Participants who reported high HIV coping self-efficacy (adjusted odds ratio (AOR) 1.86: 95% confidence interval CI 1.24-2.78, p-value=0.003), some perceived risk of HIV (AOR 2.51: 95% (CI) 1.57-4.01, p-value=0.000), and low HIV testing stigma (AOR 1.71: 95% CI 1.05-2.79, p-value=0.032) were more likely to test for HIV. Correlates of HIV coping self-efficacy included external and internal HIV stigma (AOR 1.28: 95% CI 1.25-1.32, p-value=0.000 and AOR 1.76: 95% CI 1.34-2.30, p-value=0.000, respectively), social support (AOR 2.09: 95% CI 1.19-3.68, p-value=0.010), and HIV knowledge (AOR 2.33: 95% CI 1.04-5.22, p-value=0.040). Policy and programs should focus on the interrelationships of these constructs to increase participation in HIV testing in prison. PMID:20390514

  7. Test and evaluation procedures for Sandia's Teraflops Operating System (TOS) on Janus.

    Energy Technology Data Exchange (ETDEWEB)

    Barnette, Daniel Wayne

    2005-10-01

    This report describes the test and evaluation methods by which the Teraflops Operating System, or TOS, that resides on Sandia's massively-parallel computer Janus is verified for production release. Also discussed are methods used to build TOS before testing and evaluating, miscellaneous utility scripts, a sample test plan, and a proposed post-test method for quickly examining the large number of test results. The purpose of the report is threefold: (1) to provide a guide to T&E procedures, (2) to aid and guide others who will run T&E procedures on the new ASCI Red Storm machine, and (3) to document some of the history of evaluation and testing of TOS. This report is not intended to serve as an exhaustive manual for testers to conduct T&E procedures.

  8. Rein tension acceptance in young horses in a voluntary test situation

    DEFF Research Database (Denmark)

    Christensen, Janne Winther; zharkikh, T.L.; Antoine, A.;

    2011-01-01

    ) to obtain a food reward in a voluntary test situation. On each test day, each horse was exposed to 2 control sessions (loose reins), an intermediate and a short rein session (1 min/session). Rein tension, heart rate and behaviour were recorded. Results: The horses applied significantly more tension...... on the first day (mean rein tension: 10.2 N), compared to the second and third test day (Day 2: 6.0 and Day 3: 5.7 N). The horses showed significantly more conflict behaviour in the short rein treatment. There was no treatment effect on heart rate. Conclusions and potential relevance: The horses applied...... the highest rein tension on the first day, and apparently learned to avoid the tension, rather than habituate to it. Rein tension correlated with expression of conflict behaviour, indicating that the horses found the tension aversive. Further studies should focus on the correlation between rein tension...

  9. Client characteristics and acceptability of a home-based HIV counselling and testing intervention in rural South Africa

    Directory of Open Access Journals (Sweden)

    Naik Reshma

    2012-09-01

    Full Text Available Abstract Background HIV counselling and testing (HCT is a critical gateway for addressing HIV prevention and linking people to treatment, care, and support. Since national testing rates are often less than optimal, there is growing interest in expanding testing coverage through the implementation of innovative models such as home-based HIV counselling and testing (HBHCT. With the aim of informing scale up, this paper discusses client characteristics and acceptability of an HBHCT intervention implemented in rural South Africa. Methods Trained lay counsellors offered door-to-door rapid HIV testing in a rural sub-district of KwaZulu-Natal, South Africa. Household and client data were captured on cellular phones and transmitted to a web-based data management system. Descriptive analysis was undertaken to examine client characteristics, testing history, HBHCT uptake, and reasons for refusal. Chi-square tests were performed to assess the association between client characteristics and uptake. Results Lay counsellors visited 3,328 households and tested 75% (5,086 of the 6,757 people met. The majority of testers (73.7% were female, and 57% had never previously tested. With regard to marital status, 1,916 (37.7%, 2,123 (41.7%, and 818 (16.1% were single, married, and widowed, respectively. Testers ranged in age from 14 to 98 years, with a median of 37 years. Two hundred and twenty-nine couples received couples counselling and testing; 87.8%, 4.8%, and 7.4% were concordant negative, concordant positive, and discordant, respectively. There were significant differences in characteristics between testers and non-testers as well as between male and female testers. The most common reasons for not testing were: not being ready/feeling scared/needing to think about it (34.1%; knowing his/her status (22.6%, being HIV-positive (18.5%, and not feeling at risk of having or acquiring HIV (10.1%. The distribution of reasons for refusal differed significantly by gender

  10. EFAM GTP 02 - the GKSS test procedure for determining the fracture behaviour of materials

    International Nuclear Information System (INIS)

    This document describes a unified fracture mechanics test method in procedural form for quasi-static testing of materials. It is based on the ESIS Procedures P1 and P2 and introduces additional features, such as middle cracked tension specimens, shallow cracks, the δ5 crack tip opening displacement, the crack tip opening angle, the rate of dissipated energy, testing of weldments, and guidance for statistical treatment of scatter. Special validity criteria are given for tests on specimens with low constraint. This document represents an updated version of EFAM GTP 94. (orig.)

  11. Animal testing, 3R models and regulatory acceptance : Technology transition in a risk-averse context

    NARCIS (Netherlands)

    Schiffelers, M.J.W.A.

    2016-01-01

    Risk avoidance has resulted in a broad range of regulations to guarantee the safety of products such as pharmaceuticals and chemicals. Many of these regulations rely on animal tests. About 3 million laboratory animals are used annually in Europe to meet such regulatory requirements.Regulatory animal

  12. Hanford immobilized LAW product acceptance: Initial Tanks Focus Area testing data package

    Energy Technology Data Exchange (ETDEWEB)

    JD Vienna; A Jiricka; BP McGrail; BM Jorgensen; DE Smith; BR Allen; JC Marra; DK Peeler; KG Brown; IA Reamer; WL Ebert

    2000-03-08

    The Hanford Site's mission has been to produce nuclear materials for the US Department of Energy (DOE) and its predecessors. A large inventory of radioactive and mixed waste, largely generated during plutonium production, exists in 177 underground single- and double-shell tanks. These wastes are to be retrieved and separated into low-activity waste (LAW) and high-level waste (HLW) fractions. The total volume of LAW requiring immobilization will include the LAW separated from the tank waste, as well as new wastes generated by the retrieval, pretreatment, and immobilization processes. Per the Tri-Party Agreement (1994), both the LAW and HLW will be vitrified. It has been estimated that vitrification of the LAW waste will result in over 500,000 metric tons or 200,000 m{sup 3} of immobilized LAW (ILAW) glass. The ILAW glass is to be disposed of onsite in a near-surface burial facility. It must be demonstrated that the disposal system will adequately retain the radionuclides and prevent contamination of the surrounding environment. This report describes a study of the impacts of systematic glass-composition variation on the responses from accelerated laboratory corrosion tests of representative LAW glasses. A combination of two tests, the product consistency test and vapor-hydration test, is being used to give indictations of the relative rate at which a glass could be expected to corrode in the burial scenario.

  13. Measuring Japanese EFL Student Perceptions of Internet-Based Tests with the Technology Acceptance Model

    Science.gov (United States)

    Dizon, Gilbert

    2016-01-01

    The Internet has made it possible for teachers to administer online assessments with affordability and ease. However, little is known about Japanese English as a Foreign Language (EFL) students' attitudes of internet-based tests (IBTs). Therefore, this study aimed to measure the perceptions of IBTs among Japanese English language learners with the…

  14. Review of Test Procedure for Determining HSPFs of Residential Variable-Speed Heat Pumps

    Energy Technology Data Exchange (ETDEWEB)

    Rice, C. Keith [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Munk, Jeffrey D. [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States); Shrestha, Som S. [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States)

    2015-08-01

    This report reviews the suitability of the existing Heating Seasonal Performance Factor (HSPF) ratings and testing requirements for the current generation of variable-speed (VS) air-source heat pumps. Recent field test results indicate larger discrepancies between rated HSPF and field-observed HSPF for VS models than for single-speed models in the same houses. These findings suggest that the heating season test and ratings procedure should be revisited for VS heat pumps. The ratings and testing procedures are described in ANSI/AHRI 210/240 (2008) for single-speed, two-capacity, and variable-speed units. Analysis of manufacturer and independent test performance data on VS units reveals why the current VS testing/ratings procedure results in overly optimistic HSPF ratings for some VS units relative to other types of heat pumps. This is due to a combination of extrapolation of low speed test data beyond the originally anticipated ambient temperature operating range and the constraints of unit controls, which prevent low speed operation over the range of ambient temperatures assumed in the procedure for low speed. As a result, the HSPFs of such units are being overpredicted relative to those for single- and two-capacity designs. This overprediction has been found to be significantly reduced by use in the HSPF ratings procedure of an alternative higher-load heating load line, described in a companion report (Rice et al., 2015).

  15. Reducing System Testing Effort by Focusing on Commonalities in Test Procedures

    OpenAIRE

    Hauptmann, Benedikt

    2016-01-01

    System testing is an important part of software development; however, it is also a big effort driver. High costs arise while creating, executing and maintaining test cases. This thesis proposes three ways to reduce unnecessary effort within these activities: Firstly, we help balancing between manual testing and test automation. Secondly, we support test engineers to extract clones from test cases, which makes them easier to maintain. Thirdly, we introduce Natural Language Test Smells, a techn...

  16. Acceptance Testing of a Satellite SCADA Photovoltaic-Diesel Hybrid System

    Science.gov (United States)

    Kalu, A.; Emrich, C.; Ventre, G.; Wilson, W.; Acosta, Roberto (Technical Monitor)

    2000-01-01

    Satellite Supervisory Control and Data Acquisition (SCADA) of a Photovoltaic (PV)/diesel hybrid system was tested using NASA's Advanced Communication Technology Satellite (ACTS) and Ultra Small Aperture Terminal (USAT) ground stations. The setup consisted of a custom-designed PV/diesel hybrid system, located at the Florida Solar Energy Center (FSEC), which was controlled and monitored at a "remote" hub via Ka-band satellite link connecting two 1/4 Watt USATs in a SCADA arrangement. The robustness of the communications link was tested for remote monitoring of the health and performance of a PV/diesel hybrid system, and for investigating load control and battery charging strategies to maximize battery capacity and lifetime, and minimize loss of critical load probability. Baseline hardware performance test results demonstrated that continuous two-second data transfers can be accomplished under clear sky conditions with an error rate of less than 1%. The delay introduced by the satellite (1/4 sec) was transparent to synchronization of satellite modem as well as to the PV/diesel-hybrid computer. End-to-end communications link recovery times were less than 36 seconds for loss of power and less than one second for loss of link. The system recovered by resuming operation without any manual intervention, which is important since the 4 dB margin is not sufficient to prevent loss of the satellite link during moderate to heavy rain. Hybrid operations during loss of communications link continued seamlessly but real-time monitoring was interrupted. For this sub-tropical region, the estimated amount of time that the signal fade will exceed the 4 dB margin is about 10%. These results suggest that data rates of 4800 bps and a link margin of 4 dB with a 1/4 Watt transmitter are sufficient for end-to-end operation in this SCADA application.

  17. Same, same but different. A comparative perspective on participation and acceptance in siting procedures for HLW repositories in France, Sweden and Finland

    Energy Technology Data Exchange (ETDEWEB)

    Di Nucci, Maria Rosaria; Isidoro Losada, Ana Maria; Brunnengraeber, Achim [Freie Univ. Berlin (Germany). Environmental Policy Research Centre

    2015-07-01

    This paper compares national approaches in Finland, Sweden, and France. These three EURATOM Member States are in an advanced stage of siting deep geological disposal (DGD) facilities. The procedures in these countries are largely based on voluntarism, but differ in their approach to public consultation as they were based on the so-called staged volunteer process leading to working partnership arrangements between the operator and the hosting communities and veto rights (Sweden), decisions with strong local community support and veto rights for municipalities until the final decision (Finland), and final top-down decision making after consultative processes (debat public) with the affected communities without veto rights (France). This presentation focuses on participation and acceptance issues; it analyses the different modes of governance with diverse conditions regarding transparency, trust, communication, and participation that have been at work. Moving beyond the fact that variance exists with regard to the relevant national institutional, legal, cultural, industrial, and energy frameworks, the authors take evidence from national case studies and look for common patterns.

  18. A NONPARAMETRIC PROCEDURE OF THE SAMPLE SIZE DETERMINATION FOR SURVIVAL RATE TEST

    Institute of Scientific and Technical Information of China (English)

    2000-01-01

    Objective This paper proposes a nonparametric procedure of the sample size determination for survival rate test. Methods Using the classical asymptotic normal procedure yields the required homogenetic effective sample size and using the inverse operation with the prespecified value of the survival function of censoring times yields the required sample size. Results It is matched with the rate test for censored data, does not involve survival distributions, and reduces to its classical counterpart when there is no censoring. The observed power of the test coincides with the prescribed power under usual clinical conditions. Conclusion It can be used for planning survival studies of chronic diseases.

  19. Initial land reclamation procedures related to possible Pu-cleanup activities at the Tonopah Test Range

    International Nuclear Information System (INIS)

    If areas of the Tonopah Test Range (TTR) are to be used for experimental tests of procedures for clean-up of 239Pu contamination, there are experiences in the Great Basin Desert portions of the Nevada Test Site (NTS) which can serve as guides to reclamation and revegetation of such arid lands. Procedures which will encourage development of the grasses Hilaria jamesii and Oryzopsis hymenoides, as well as the perennial shrubs Eurotia lanata and Atriplex canescens would greatly improve the area as range land

  20. Kilowatt Isotope Power System: test procedure for the EHSA multifoil insulation. 77-KIPS-109

    Energy Technology Data Exchange (ETDEWEB)

    1977-01-01

    The procedure is described for conducting the multifoil insulation meltdown test on the electrical heat source assembly (EHSA). The results of the tests will be used to verify the predicted sequence of events during multifoil insulation meltdown and to determine the ultimate steady-state temperatures of the various EHSA components.

  1. 40 CFR 92.125 - Pre-test procedures and preconditioning.

    Science.gov (United States)

    2010-07-01

    ... valve is measured using in-use pressures and bypass flows (see § 92.118). (C) The response time measured....125 Pre-test procedures and preconditioning. (a) Locomotive testing. (1) Determine engine lubricating... fuel supply system and purge as necessary; determine that the fuel to be used during emission...

  2. 75 FR 42612 - Energy Conservation Program for Consumer Products: Test Procedure for Microwave Ovens

    Science.gov (United States)

    2010-07-22

    ... consumption into the energy descriptor for that product unless the current test procedures already fully.... 62 FR 51976. That final rule incorporated portions of the International Electrotechnical Commission... proposed correcting a technical error in the calculation of microwave test cooking energy output. 73...

  3. 77 FR 8178 - Test Procedures for Central Air Conditioners and Heat Pumps: Public Meeting

    Science.gov (United States)

    2012-02-14

    ...; ] DEPARTMENT OF ENERGY 10 CFR Part 430 Test Procedures for Central Air Conditioners and Heat Pumps: Public... discuss methodologies and gather comments on testing residential central air conditioners and heat pumps... residential central air conditioners and heat pumps that are single phase with rated cooling capacities...

  4. 78 FR 19606 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Furnace Fans

    Science.gov (United States)

    2013-04-02

    ... the test procedure rulemaking for furnace fans. 77 FR 28674. In the NOPR, DOE proposed a rating metric... furnace fans energy conservation standard rulemaking on July 10, 2012. 77 FR 40530. For the preliminary... language in the proposed regulatory text that specified that furnace fans be tested in-situ. 77 FR at...

  5. 78 FR 38455 - Energy Conservation Program: Test Procedures for Electric Motors

    Science.gov (United States)

    2013-06-26

    ... test procedures under 10 CFR part 431 both for electric motors and small electric motors. 73 FR 78220. DOE finalized key provisions related to small electric motor testing in a 2009 final rule at 74 FR... at 77 FR 26608 (May 4, 2012). Today's notice of proposed rulemaking (NOPR) focuses on electric...

  6. 76 FR 57612 - Energy Efficiency Program for Consumer Products: Test Procedures for Residential Refrigerators...

    Science.gov (United States)

    2011-09-15

    ..., and freezers (collectively, ``refrigeration products''). 75 FR 78810. The amended test procedures for..., Appendix A and Appendix B, that incorporate the same revisions made to Appendix A1 and Appendix B1. 75 FR... multiple defrost cycle test that DOE presented in its Interim Final Rule. See 76 FR 16760 (March 25,...

  7. 77 FR 68784 - Standard Test Procedures Approval Process for Respirators To Be Used in Wildland Fire-Fighting...

    Science.gov (United States)

    2012-11-16

    ... HUMAN SERVICES Centers for Disease Control and Prevention Standard Test Procedures Approval Process for Respirators To Be Used in Wildland Fire-Fighting Operations; Standard Test Procedures for Composite Multi-Gas.... NIOSH has developed test procedures for a composite particulate and multi-gas protection for APR...

  8. Evaluation and Acceptability of a Simplified Test of Visual Function at Birth in a Limited-Resource Setting.

    Science.gov (United States)

    Carrara, Verena I; Darakomon, Mue Chae; Thin, Nant War War; Paw, Naw Ta Kaw; Wah, Naw; Wah, Hser Gay; Helen, Naw; Keereecharoen, Suporn; Paw, Naw Ta Mlar; Jittamala, Podjanee; Nosten, François H; Ricci, Daniela; McGready, Rose

    2016-01-01

    Neurological examination, including visual fixation and tracking of a target, is routinely performed in the Shoklo Malaria Research Unit postnatal care units on the Thailand-Myanmar border. We aimed to evaluate a simple visual newborn test developed in Italy and performed by non-specialized personnel working in neonatal care units. An intensive training of local health staff in Thailand was conducted prior to performing assessments at 24, 48 and 72 hours of life in healthy, low-risk term singletons. The 48 and 72 hours results were then compared to values obtained to those from Italy. Parents and staff administering the test reported on acceptability. One hundred and seventy nine newborns, between June 2011 and October 2012, participated in the study. The test was rapidly completed if the infant remained in an optimal behavioral stage (7 ± 2 minutes) but the test duration increased significantly (12 ± 4 minutes, p resource setting have a visual response similar to that obtained to gestational age matched newborns in Italy. It is possible to use these results as a reference set of values for the visual assessment in Karen and Burmese infants in the first 72 hours of life. The utility of the 24 hours test should be pursued. PMID:27300137

  9. Test procedures and instructions for single shell tank saltcake cesium removal with crystalline silicotitanate

    Energy Technology Data Exchange (ETDEWEB)

    Duncan, J.B.

    1997-01-07

    This document provides specific test procedures and instructions to implement the test plan for the preparation and conduct of a cesium removal test, using Hanford Single Shell Tank Saltcake from tanks 24 t -BY- I 10, 24 1 -U- 108, 24 1 -U- 109, 24 1 -A- I 0 1, and 24 t - S-102, in a bench-scale column. The cesium sorbent to be tested is crystalline siticotitanate. The test plan for which this provides instructions is WHC-SD-RE-TP-024, Hanford Single Shell Tank Saltcake Cesium Removal Test Plan.

  10. Storage, transportation, test marketing and consumer acceptance of irradiated potatoes, onions, garlic and fruit

    International Nuclear Information System (INIS)

    The Shanghai Irradiation Center irradiated 830 tonnes of garlic 395 tonnes of potatoes 620 tonnes of onions and 50 tonnes of apples from 1990 to 1993 for test marketing. A consumer trial on irradiated apples was conducted. The results showed that 90% of consumers liked irradiated fruit and would be willing to buy them again. SIC established technological standards for the irradiation of garlic, potatoes and onions based on the codes of good irradiation practices recommended by the International Consultative Group on Food Irradiation. These standards were followed during the conduct of the research studies reported in this paper. A cost-benefit analysis of irradiated apples and garlic was performed. It was observed that irradiated apples would provide a profit of 52% and irradiated garlic 100% compared to unirradiated products. Data from an economic benefits study showed that products requiring a low absorbed dose have a higher profit margin. Therefore, the irradiation of garlic, potatoes and apples would be more profitable than radiation treatment of medical supplies and industrial products. (author). 7 refs, 3 figs, 10 tabs

  11. Development of interim test methods and procedures for determining the performance of small photovoltaic systems

    Energy Technology Data Exchange (ETDEWEB)

    McNutt, P.; Kroposki, B.; Hansen, R.; Algra, K.; DeBlasio, R. [National Renewable Energy Lab., Golden, CO (United States)

    1998-09-01

    The National Renewable Energy Laboratory (NREL) is developing tests and procedures that will determine if the configuration of a small photovoltaic (PV) system is suitable for its intended use, and if the system will perform as specified. An overview of these procedures is presented in this paper. Development of standard test procedures will allow designers, manufacturers, system integrators, users, and independent laboratories to assess the performance of PV systems under outdoor prevailing conditions. An NREL Technical Report detailing the procedures is under way, and the IEEE Standards Coordinating Committee 21 (SCC21) has established a project on this subject. The work will be submitted to the IEEE SCC21 and International Electrotechnical Commission Technical Committee 82 (IEC TC82) for consideration as a consensus standard. Certification bodies such as PowerMark and PV Global Approval Program (PVGAP) may adopt the IEC and IEEE documents when testing systems. Developing standardized test methods and procedures at NREL to evaluate the outdoor performance of PV systems will encourage product quality and promote PV standards development. Standardized tests will assure people that PV systems will perform as specified for their intended applications. As confidence in PV systems increases, the successful commercialization of PV will grow internationally.

  12. HIV testing in non-traditional settings--the HINTS study: a multi-centre observational study of feasibility and acceptability.

    Directory of Open Access Journals (Sweden)

    Michael Rayment

    Full Text Available BACKGROUND: UK guidelines recommend routine HIV testing in healthcare settings if the local diagnosed HIV prevalence >2/1000 persons. This prospective study assessed the feasibility and acceptability, to patients and staff, of routinely offering HIV tests in four settings: Emergency Department, Acute Care Unit, Dermatology Outpatients and Primary Care. Modelling suggested the estimated prevalence of undiagnosed HIV infection in attendees would exceed 1/1000 persons. The prevalence identified prospectively was not a primary outcome. METHODS: Permanent staff completed questionnaires assessing attitudes towards routine HIV testing in their workplace before testing began. Subsequently, over a three-month period, patients aged 16-65 were offered an HIV test by study staff. Demographics, uptake, results, and departmental activity were collected. Subsets of patients completed questionnaires. Analyses were conducted to identify factors associated with test uptake. FINDINGS: Questionnaires were received from 144 staff. 96% supported the expansion of HIV testing, but only 54% stated that they would feel comfortable delivering testing themselves, with 72% identifying a need for training. Of 6194 patients offered a test, 4105 (66·8% accepted (61·8-75·4% across sites. Eight individuals were diagnosed with HIV (0-10/1000 across sites and all transferred to care. Younger people, and males, were more likely to accept an HIV test. No significant associations were found between uptake and ethnicity, or clinical site. Questionnaires were returned from 1003 patients. The offer of an HIV test was acceptable to 92%. Of respondents, individuals who had never tested for HIV before were more likely to accept a test, but no association was found between test uptake and sexual orientation. CONCLUSIONS: HIV testing in these settings is acceptable, and operationally feasible. The strategy successfully identified, and transferred to care, HIV-positive individuals. However

  13. Results of the Updated NASA Kennedy Space Center 50-MHz Doppler Radar Wind Profiler Operational Acceptance Test

    Science.gov (United States)

    Barbre', Robert E., Jr.; Deker, Ryan K.; Leahy, Frank B.; Huddleston, Lisa

    2016-01-01

    We present here the methodology and results of the Operational Acceptance Test (OAT) performed on the new Kennedy Space Center (KSC) 50-MHz Doppler Radar Wind Profiler (DRWP). On day-of-launch (DOL), space launch vehicle operators have used data from the DRWP to invalidate winds in prelaunch loads and trajectory assessments due to the DRWP's capability to quickly identify changes in the wind profile within a rapidly-changing wind environment. The previous DRWP has been replaced with a completely new system, which needs to undergo certification testing before being accepted for use in range operations. The new DRWP replaces the previous three-beam system made of coaxial cables and a copper wire ground plane with a four-beam system that uses Yagi antennae with enhanced beam steering capability. In addition, the new system contains updated user interface software while maintaining the same general capability as the previous system. The new DRWP continues to use the Median Filter First Guess (MFFG) algorithm to generate a wind profile from Doppler spectra at each range gate. DeTect (2015) contains further details on the upgrade. The OAT is a short-term test designed so that end users can utilize the new DRWP in a similar manner to the previous DRWP during mission operations at the Eastern Range in the midst of a long-term certification process. This paper describes the Marshall Space Flight Center Natural Environments Branch's (MSFC NE's) analyses to verify the quality and accuracy of the DRWP's meteorological data output as compared to the previous DRWP. Ultimately, each launch vehicle program has the responsibility to certify the system for their own use.

  14. Flaw acceptance criteria taking into consideration the NDT: radiographic and ultrasonic testing. Analysis through the fracture mechanics methods

    International Nuclear Information System (INIS)

    The present study compares and evaluates the flaw acceptance criteria of the non-destructive inspections meeting European Community standards, through the application of the fracture mechanics methods that were determined and verified by the previous activity. Some choices were made; these, however, do not change the general validity of the conclusions. Shaved full-penetration butt welds of Class 1 components making up the primary circuit were considered and the following parameters varied: standards: French, German, Italian (ASME III) and UK; material: AISI 316 and low alloy steel A 533; base material and weld metal; temperature: RT, 370 deg C for the austenitic and 260 deg C for the ferritic steel; ultrasonic and radiographic methods; defect position: surface and internal; stress condition: situations with different primary and secondary stresses. From a preliminary examination of this study it is evident that the large quantity of results available and the abundance of information contained therein make a simple and exhaustive synthesis difficult. In fact, different analyses are possible and we have, therefore, limited the research to activities to perform a comparison and a general evaluation of the acceptance criteria of the non-destructive testing. (authors). 57 refs., 25 figs., 11 tabs

  15. Several submaximal exercise tests are reliable, valid and acceptable in people with chronic pain, fibromyalgia or chronic fatigue: a systematic review

    Directory of Open Access Journals (Sweden)

    Julia Ratter

    2014-09-01

    [Ratter J, Radlinger L, Lucas C (2014 Several submaximal exercise tests are reliable, valid and acceptable in people with chronic pain, fibromyalgia or chronic fatigue: a systematic review. Journal of Physiotherapy 60: 144–150

  16. Safety testing and operational procedures for self-developed radiofrequency coils.

    Science.gov (United States)

    Hoffmann, Jens; Henning, Anke; Giapitzakis, Ioannis A; Scheffler, Klaus; Shajan, G; Pohmann, Rolf; Avdievich, Nikolai I

    2016-09-01

    The development of novel radiofrequency (RF) coils for human ultrahigh-field (≥7 T), non-proton and body applications is an active field of research in many MR groups. Any RF coil must meet the strict requirements for safe application on humans with respect to mechanical and electrical safety, as well as the specific absorption rate (SAR) limits. For this purpose, regulations such as the International Electrotechnical Commission (IEC) standard for medical electrical equipment, vendor-suggested test specifications for third party coils and custom-developed test procedures exist. However, for higher frequencies and shorter wavelengths in ultrahigh-field MR, the RF fields may become extremely inhomogeneous in biological tissue and the risk of localized areas with elevated power deposition increases, which is usually not considered by existing safety testing and operational procedures. In addition, important aspects, such as risk analysis and comprehensive electrical performance and safety tests, are often neglected. In this article, we describe the guidelines used in our institution for electrical and mechanical safety tests, SAR simulation and verification, risk analysis and operational procedures, including coil documentation, user training and regular quality assurance testing, which help to recognize and eliminate safety issues during coil design and operation. Although the procedure is generally applicable to all field strengths, specific requirements with regard to SAR-related safety and electrical performance at ultrahigh-field are considered. The protocol describes an internal procedure and does not reflect consensus among a large number of research groups, but rather aims to stimulate further discussion related to minimum coil safety standards. Furthermore, it may help other research groups to establish their own procedures. Copyright © 2015 John Wiley & Sons, Ltd. PMID:25851551

  17. Hanford spent nuclear fuel cold vacuum drying proof of performance test procedure

    International Nuclear Information System (INIS)

    This document provides the test procedure for cold testing of the first article skids for the Cold Vacuum Drying (CVD) process at the Facility. The primary objective of this testing is to confirm design choices and provide data for the initial start-up parameters for the process. The current scope of testing in this document includes design verification, drying cycle determination equipment performance testing of the CVD process and MCC components, heat up and cool-down cycle determination, and thermal model validation

  18. Proton injector acceptance tests for a Dielectric Wall Accelerator (DWA):characterisation of Advanced Injection System of Light Ions (AISLI)

    Energy Technology Data Exchange (ETDEWEB)

    Peng, S.X., E-mail: sxpeng@pku.edu.cn; Ren, H.T.; Xu, Y.; Zhang, T.; Zhao, J.; Zhang, A.L.; Guo, Z.Y.; Chen, J.E.

    2014-11-01

    To demonstrate the acceleration capability of a Dielectric Wall Accelerator (DWA), a proton injector has been developed at Peking University (PKU). It is composed of a compact permanent magnet 2.45 GHz Electron Cyclotron Resonance Ion Source (PKU PMECRIS) associated with a LEBT of two electrostatic Einzel lenses [1]. This injector is named as Advanced Injector System of Light Ions (AISLI). The acceptance tests have been performed with a 40 keV–55 mA hydrogen beam successfully passing through a ϕ 10 mm aperture diaphragm. This diaphragm is located 200 mm downstream the plasma emission hole at the location of the future DWA entrance flange. The beam rms emittance reached about 0.10 π mm mrad in pulsed mode. This article describes the AISLI experimental setup, the measurement principle and the obtained beam characteristics.

  19. A test procedure for energetic and performance analysis of cold appliances for the food industry

    Science.gov (United States)

    Armani, F.; Boscolo, A.

    2013-09-01

    In this article we present a novel approach for the characterization of cold appliances and in particular of refrigerators based on the standard vapour compression cycle with a reciprocating on/off compressor. The test procedure is based on a virtual instrument that perform both the stimulus and the data acquisition on the device under test. Acquired data is elaborated to fit a semi-empirical model based on the energetic balances between thermal and electrical sub systems and the heat exchanged with the environment. This approach results in a simple method to calculate useful parameters of the refrigerator, such as energetic performance, cooling effect and limit values of thermal loads. The test procedure requires only a few temperatures and the electric power consumption to be monitored, resulting in a low impact on the refrigerator. Preliminary tests showed a good estimation of parameters and prediction of energy consumption and heat extraction capacity of the refrigerator under test.

  20. Mediator, Moderator and Intervening Variables in Marketing Researchs: Conceptualization, Differences and Statistical Procedures and Tests

    Directory of Open Access Journals (Sweden)

    shahriar Azizi

    2013-09-01

    Full Text Available Recent years marketing theories has been developed increasingly. This development needs paying attention to variables such as: moderator and mediator for extending marketing research models. Marketing and management researchers need to be informed about the meaning of mediator, moderator and intervening variables and accurate statistical procedures and tests for identification. By understanding the true meaning of mediator, moderator and intervening variables, providing more accurate marketing models that fits real word become easier. In this paper the meanings of those variables were presented and compared. In the next sections of the paper, statistical procedures and tests provided.

  1. A new test procedure for biogenic sulfuric acid corrosion of concrete

    Science.gov (United States)

    Vincke; Verstichel; Monteny; Verstraete

    1999-01-01

    A new test method is described for biogenic sulfuric acid corrosion of concrete, more specifically in sewer conditions. The aim of the new test method is the development of an accelerated and reproducible procedure for monitoring the resistance of different types of concrete with regard to biogenic sulfuric acid corrosion. This experimental procedure reflects worst case conditions by providing besides H2S, also an enrichment of thiobacilli and biologically produced sulfur. By simulating the cyclic processes occurring in sewer pipes, significant differences between concrete mixtures could be detected after 51 days. Concrete modified by a styrene-acrylic ester polymer demonstrated a higher resistance against biogenic sulfuric acid attack.

  2. Test procedures to detect a loss of material in a sequence of balance periods

    International Nuclear Information System (INIS)

    A workshop on Near-Real-Time Accountancy (NRTA) was held in KfK which came to a preliminary end in December 1982. In the framework of this workshop a number of sequential statistical test procedures were proposed which can be used in the case of a NRTA based safeguards regime. In the report the most promising test procedures are investigated. The analysis is based on the chemical separation process of a large model reprocessing facility with a throughput or 1000 tonnes per year. (orig.)

  3. Test Procedures for Characterizing, Evaluating, and Managing Separator Materials used in Secondary Alkaline Batteries

    Science.gov (United States)

    Guasp, Edwin; Manzo, Michelle A.

    1997-01-01

    Secondary alkaline batteries, such as nickel-cadmium and silver-zinc, are commonly used for aerospace applications. The uniform evaluation and comparison of separator properties for these systems is dependent upon the measurement techniques. This manual presents a series of standard test procedures that can be used to evaluate, compare, and select separator materials for use in alkaline batteries. Detailed test procedures evaluating the following characteristics are included in this manual: physical measurements of thickness and area weight, dimensional stability measurements, electrolyte retention, resistivity, permeability as measured via bubble pressure, surface evaluation via SEM, chemical stability, and tensile strength.

  4. 49 CFR Appendix B to Part 173 - Procedure for Testing Chemical Compatibility and Rate of Permeation in Plastic Packaging and...

    Science.gov (United States)

    2010-10-01

    ... 49 Transportation 2 2010-10-01 2010-10-01 false Procedure for Testing Chemical Compatibility and..., App. B Appendix B to Part 173—Procedure for Testing Chemical Compatibility and Rate of Permeation in Plastic Packaging and Receptacles 1. The purpose of this procedure is to determine the...

  5. 9 CFR 147.6 - Procedure for determining the status of flocks reacting to tests for Mycoplasma gallisepticum...

    Science.gov (United States)

    2010-01-01

    ... 9 Animals and Animal Products 1 2010-01-01 2010-01-01 false Procedure for determining the status... meleagridis. 147.6 Section 147.6 Animals and Animal Products ANIMAL AND PLANT HEALTH INSPECTION SERVICE... Blood Testing Procedures § 147.6 Procedure for determining the status of flocks reacting to tests...

  6. Efficient Noninferiority Testing Procedures for Simultaneously Assessing Sensitivity and Specificity of Two Diagnostic Tests

    Directory of Open Access Journals (Sweden)

    Guogen Shan

    2015-01-01

    Full Text Available Sensitivity and specificity are often used to assess the performance of a diagnostic test with binary outcomes. Wald-type test statistics have been proposed for testing sensitivity and specificity individually. In the presence of a gold standard, simultaneous comparison between two diagnostic tests for noninferiority of sensitivity and specificity based on an asymptotic approach has been studied by Chen et al. (2003. However, the asymptotic approach may suffer from unsatisfactory type I error control as observed from many studies, especially in small to medium sample settings. In this paper, we compare three unconditional approaches for simultaneously testing sensitivity and specificity. They are approaches based on estimation, maximization, and a combination of estimation and maximization. Although the estimation approach does not guarantee type I error, it has satisfactory performance with regard to type I error control. The other two unconditional approaches are exact. The approach based on estimation and maximization is generally more powerful than the approach based on maximization.

  7. Experimental testing procedures and dynamic model validation for vanadium redox flow battery storage system

    Science.gov (United States)

    Baccino, Francesco; Marinelli, Mattia; Nørgård, Per; Silvestro, Federico

    2014-05-01

    The paper aims at characterizing the electrochemical and thermal parameters of a 15 kW/320 kWh vanadium redox flow battery (VRB) installed in the SYSLAB test facility of the DTU Risø Campus and experimentally validating the proposed dynamic model realized in Matlab-Simulink. The adopted testing procedure consists of analyzing the voltage and current values during a power reference step-response and evaluating the relevant electrochemical parameters such as the internal resistance. The results of different tests are presented and used to define the electrical characteristics and the overall efficiency of the battery system. The test procedure has general validity and could also be used for other storage technologies. The storage model proposed and described is suitable for electrical studies and can represent a general model in terms of validity. Finally, the model simulation outputs are compared with experimental measurements during a discharge-charge sequence.

  8. Assessment of Calculation Procedures for Piles in Clay Based on Static Loading Tests

    DEFF Research Database (Denmark)

    Augustesen, Anders; Andersen, Lars

    2008-01-01

    College in London. The calculation procedures are assessed based on an established database of static loading tests. To make a consistent evaluation of the design methods, corrections related to undrained shear strength and time between pile driving and testing have been employed. The study indicates...... that the interpretation of the field tests is of paramount importance, both with regard to the soil profile and the loading conditions. Based on analyses of 253 static pile loading tests distributed on 111 sites, API-RP2A provides the better description of the data. However, it should be emphasised that some input...

  9. Do Ethical Judgments Depend on the Type of Response Scale? Comparing Acceptability versus Unacceptability Judgments in the Case of Life-Ending Procedures

    Science.gov (United States)

    Sastre, Maria Teresa Munoz; Gonzalez, Charlene; Lhermitte, Astrid; Sorum, Paul C.; Mullet, Etienne

    2010-01-01

    Using Functional Measurement (Anderson, 2008), Frileux, Lelievre, Munoz Sastre, Mullet, and Sorum (2003) examined the joint impact of several key factors on lay people's judgments of the acceptability of physicians' interventions to end patients' lives. The level of acceptability was high, and the information integration rule that best described…

  10. 77 FR 58804 - Testing of Product Samples for Listeria monocytogenes: Changes in Procedures

    Science.gov (United States)

    2012-09-24

    ... Poultry Products'' (68 FR 34208; Jun. 6, 2003), which declares that post-lethality exposed RTE products... Food Safety and Inspection Service Testing of Product Samples for Listeria monocytogenes: Changes in... Inspection Service (FSIS) is announcing changes in procedures for Listeria (L.) monocytogenes...

  11. 76 FR 971 - Energy Conservation Program for Consumer Products: Test Procedures for Clothes Dryers and Room...

    Science.gov (United States)

    2011-01-06

    ... editorial and minor technical changes. 46 FR 27324. The existing clothes dryer test procedure incorporates... rulemakings concurrent with the clothes dryer rulemaking. 72 FR 57254 (October 9, 2007). As stated above, DOE... dryer after the load reaches an RMC of 5 percent. 75 FR 37594, 37599 (June 29, 2010). DOE...

  12. 75 FR 14287 - Energy Conservation Program: Test Procedures for Fluorescent Lamp Ballasts

    Science.gov (United States)

    2010-03-24

    ... test procedure (appendix Q) can be shown to have a measurement uncertainty on the order of 5 percent... appearing on page 2. DOE understands NEMA's concerns regarding the measurement uncertainty related to the... impacted by measurement ] uncertainty because much of the T8 ballast market is high-frequency...

  13. 75 FR 31223 - Energy Conservation Program for Consumer Products: Test Procedure for Residential Central Air...

    Science.gov (United States)

    2010-06-02

    ... External Static Pressure Values 5. Fan Time Delay Relays 6. Inverter-Driven Compressors 7. Addition of... procedure to convert to wet-coil cyclic testing. 6. Inverter-Driven Compressors Mitsubishi Electric and Electronics USA, Inc. (MEUS) commented that new systems incorporating inverter-driven compressor...

  14. 16 CFR 1610.35 - Procedures for testing special types of textile fabrics under the standard.

    Science.gov (United States)

    2010-01-01

    ... textile fabrics under the standard. 1610.35 Section 1610.35 Commercial Practices CONSUMER PRODUCT SAFETY COMMISSION FLAMMABLE FABRICS ACT REGULATIONS STANDARD FOR THE FLAMMABILITY OF CLOTHING TEXTILES Rules and Regulations § 1610.35 Procedures for testing special types of textile fabrics under the standard. (a)...

  15. 77 FR 33106 - Energy Conservation Program: Test Procedure for Microwave Ovens

    Science.gov (United States)

    2012-06-05

    ... energy use for microwave ovens (73 FR 62134), interested parties commented that pure water has relatively... consistent water content. In addition, Russet potatoes were identified to be the most likely to be available... evaluate potential test procedure amendments to provide methods of measuring energy use for microwave...

  16. 77 FR 28805 - Energy Conservation Program: Test Procedures for Microwave Ovens

    Science.gov (United States)

    2012-05-16

    ... covered product for the DOE microwave oven test procedure. 76 FR 72332, 72336 (Nov. 23, 2011). DOE... cooking product. 76 FR 72332, 72336 (Nov. 23, 2011). \\4\\ In previous stages of this rulemaking, DOE... consumption, or other energy descriptor for each covered product, unless the Secretary determines that--...

  17. 76 FR 56339 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Furnaces and...

    Science.gov (United States)

    2011-09-13

    ... link to the docket web page can be found at: http://www.regulations.gov/#!docketDetail ;dct=FR%252BPR... efficiency, energy consumption, or other energy descriptor for each covered product, unless the Secretary... EISA 2007. 75 FR 64621 (Oct. 20, 2010). For a more detailed procedural history of the test...

  18. 77 FR 65941 - Energy Conservation Program: Test Procedures for Residential Dishwashers, Dehumidifiers, and...

    Science.gov (United States)

    2012-10-31

    ... product energy conservation standards addressing standby mode and off mode energy use. 77 FR 31444, 31451... energy efficiency, energy consumption, or other energy descriptor for each covered product, unless the... FR 39964 (Aug. 8, 1977). Since that time, the dishwasher test procedure has undergone a number...

  19. 75 FR 10949 - Energy Conservation Program for Certain Commercial and Industrial Equipment: Test Procedure for...

    Science.gov (United States)

    2010-03-09

    ..., or other energy descriptor for each covered product for which DOE's current test procedures do not... established by EISA 2007 does not apply to all ballasts. 74 Federal Register (FR) 33171, 33174 (July 10, 2009... ballasts, and tentatively concluded that off mode is not applicable. 74 FR 33171, 33172-73 (July 10,...

  20. 77 FR 76831 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Furnaces and...

    Science.gov (United States)

    2012-12-31

    ... page can be found at: http://www.regulations.gov/#!docketDetail ;dct=FR%252BPR%252BN%252BO%252BSR;rpp... descriptor for each covered product, unless the Secretary determines that: (1) The current test procedures... mode and off mode energy consumption of these products, as required by EISA 2007. 75 FR 64621. For...

  1. Which personality dimensions do puppy tests measure? A systematic procedure for categorizing behavioral assays.

    Science.gov (United States)

    McGarrity, Monica E; Sinn, David L; Gosling, Samuel D

    2015-01-01

    With the recent increase in interest in personality in dogs, behavioral assays of their behavior have proliferated. There has been particularly strong interest in predicting adult behavior from puppy tests. As a result, researchers and practitioners seeking to measure personality in puppies are faced with a bewildering array of options and no clear guide as to what behavioral assays have been developed or which personality dimensions those assays measure effectively. To address this issue, we used an 'expert-categorization' procedure-a standardized method often used in the course of meta-analyses-to identify the subset of those assays consensually judged to measure major personality dimensions effectively. We used this procedure to identify all relevant puppy tests and to categorize them in terms of their ability to measure nine personality dimensions identified in previous research (activity, aggressiveness, boldness/self-assuredness, exploration, fearfulness/nervousness, reactivity, sociability, submissiveness, trainability/responsiveness). Specifically, we identified 264 assay subtests, derived from 47 studies, which were subjected to a standardized categorization procedure undertaken independently by six expert judges. These procedures yielded a set of behavioral tests judged to measure the nine dimensions effectively and also demonstrated a widely applicable method for developing and evaluating behavioral test batteries. This article is part of a Special Issue entitled: Canine Behavior. PMID:25267465

  2. 78 FR 42719 - Test Procedures for Showerheads, Faucets, Water Closets, Urinals, and Commercial Prerinse Spray...

    Science.gov (United States)

    2013-07-17

    ... faucets.\\3\\ 63 FR 13308 (March 18, 1998). The current test procedures for faucets and showerheads, as...-2011. 77 FR 31742, 31744. DOE requested comments and information on prospective methods for verifying... force requirement or, alternatively, make verification unnecessary. 77 FR at 31746-31748. \\4\\...

  3. Testing taste sensitivity and aversion in very young children : development of a procedure

    NARCIS (Netherlands)

    Visser, J; Kroeze, J.H.A.; Kamps, WA; Bijleveld, CMA

    2000-01-01

    Taste perception in 45 3- to 6-year-old children was tested using procedures specifically designed for this age group. Detection thresholds for sucrose and urea were measured by a staircase method and aversion to urea was assessed hedonically, using drawings of facial expressions. All children under

  4. Testing taste sensitivity and aversion in very young children: development of a procedure

    NARCIS (Netherlands)

    Visser, J.; Kroeze, J.H.A.; Kamps, W.A.; Bijleveld, C.M.A.

    2000-01-01

    Taste perception in 45 3- to 6-year-old children was tested using procedures specifically designed for this age group. Detection thresholds for sucrose and urea were measured by a staircase method and aversion to urea was assessed hedonically, using drawings of facial expressions. All children under

  5. 40 CFR 63.1282 - Test methods, compliance procedures, and compliance demonstrations.

    Science.gov (United States)

    2010-07-01

    ... performance criteria of Method 21, 40 CFR part 60, appendix A, except the instrument response factor criteria... requirements of 40 CFR part 265, subpart O. (vi) A control device for which a performance test was conducted... demonstrations. (a) Determination of glycol dehydration unit flowrate or benzene emissions. The procedures...

  6. 76 FR 47178 - Energy Efficiency Program: Test Procedure for Lighting Systems (Luminaires)

    Science.gov (United States)

    2011-08-04

    ... 1904-AC50 Energy Efficiency Program: Test Procedure for Lighting Systems (Luminaires) AGENCY: Office of Energy Efficiency and Renewable Energy, Department of Energy. ACTION: Request for Information. SUMMARY... Energy Efficiency and Renewable Energy, Building Technologies Program, EE-2J, 1000 Independence...

  7. 76 FR 56347 - Energy Conservation Program for Consumer Products: Test Procedures for Residential Water Heaters...

    Science.gov (United States)

    2011-09-13

    ... Part 430 RIN 1904-AB95 Energy Conservation Program for Consumer Products: Test Procedures for Residential Water Heaters, Direct Heating Equipment, and Pool Heaters (Standby Mode and Off Mode) AGENCY... residential water heaters, direct heating equipment, and pool heaters to include provisions for...

  8. 78 FR 79637 - Energy Conservation Program: Test Procedure for Set-Top Boxes

    Science.gov (United States)

    2013-12-31

    ... determined that STBs and network equipment qualify as a covered product. 76 FR at 34914. Subsequently, DOE.... 76 FR at 78174. DOE then published a notice of proposed rulemaking (NOPR) on January 23, 2013 to establish a new test procedure focused exclusively on STBs. 78 FR 5076. DOE held a public meeting...

  9. 75 FR 71596 - Energy Efficiency Program for Certain Commercial and Industrial Equipment: Test Procedures for...

    Science.gov (United States)

    2010-11-24

    ... equipment classes on January 9, 2009. 74 FR 1091. Manufacturers of covered equipment, including commercial...-2004) for measuring refrigerated compartment volume. 71 FR 71370. These industry standards for... the DOE test procedure for commercial refrigeration equipment. 71 FR 71370; 10 CFR 431.63-431.64....

  10. 78 FR 44189 - Petition for Modification of Single Car Air Brake Test Procedures

    Science.gov (United States)

    2013-07-23

    ..., 2000 (65 FR 19477). Robert C. Lauby, Deputy Associate Administrator for Regulatory and Legislative... Federal Railroad Administration Petition for Modification of Single Car Air Brake Test Procedures In... the Federal Railroad Administration (FRA) per 49 CFR 232.307 to modify the single car air brake...

  11. A complementary emissions test for light-duty vehicles: Assessing the technical feasibility of candidate procedures

    OpenAIRE

    WEISS MARTIN; Bonnel, Pierre; HUMMEL RUDOLF; STEININGER Nikolaus

    2012-01-01

    Light-duty diesel vehicles emit on the road substantially more nitrogen oxides than permitted by regulatory emissions standards. The European Commission addresses this problem by developing a complementary emissions test procedure for the type approval and in-service conformity testing of these vehicles. To facilitate the technical development, the European Commission established in January 2011 the Real-Driving Emissions - Light-Duty Vehicles (RDE-LDV) working group that is open to all stake...

  12. 16 CFR Table 3 to Part 1512 - Minimum Acceptable Values for the Quantity A Defined in the Retroreflective Tire and Rim Test...

    Science.gov (United States)

    2010-01-01

    ... Defined in the Retroreflective Tire and Rim Test Procedure 3 Table 3 to Part 1512 Commercial Practices CONSUMER PRODUCT SAFETY COMMISSION FEDERAL HAZARDOUS SUBSTANCES ACT REGULATIONS REQUIREMENTS FOR BICYCLES... Retroreflective Tire and Rim Test Procedure Observation angle (degrees) Entrance angle (degrees)...

  13. Acceptability and feasibility of using established geosocial and sexual networking mobile applications to promote HIV and STD testing among men who have sex with men

    OpenAIRE

    Sun, Christina J.; Stowers, Jason; Miller, Cindy; BACHMANN, LAURA H.; Rhodes, Scott D.

    2015-01-01

    This study is the first published multi-app study, of which we are aware, to evaluate both the acceptability and feasibility of providing sexual health information and HIV/STD testing referrals via established geosocial and sexual networking apps for MSM. Data were collected using an online survey and through four apps (A4A Radar, Grindr, Jack’d, and Scruff). Two-thirds (64%) found apps to be an acceptable source for sexual health information. MSM who found apps as acceptable were more likely...

  14. SLURRY MIX EVAPORATOR BATCH ACCEPTABILITY AND TEST CASES OF THE PRODUCT COMPOSITION CONTROL SYSTEM WITH THORIUM AS A REPORTABLE ELEMENT

    Energy Technology Data Exchange (ETDEWEB)

    Edwards, T.

    2010-10-07

    (SRNL) was requested to conduct. SRNL issued a Task Technical and Quality Assurance (TT&QA) plan [4] in response to the SRR request. The conclusions provided in this report are that no changes need to be made to the SME acceptability process (i.e., no modifications to WSRC-TR-95-00364, Revision 5, are needed) and no changes need to be made to the Product Composition Control System (PCCS) itself (i.e. the spreadsheet utilized by Waste Solidification Engineering (WSE) for acceptability decisions does not require modification) in response to thorium becoming a reportable element for DWPF operations. In addition, the inputs and results for the two test cases requested by WSE for use in confirming the successful activation of thorium as a reportable element for DWPF operations during the processing of SB6 are presented in this report.

  15. Development of helium leak testing system and procedure for testing welds of steam generator

    International Nuclear Information System (INIS)

    FBR steam generator (SG) is a vertical shell and tube type heat exchanger with sodium on shell side and water/steam on tube side. As sodium and water are highly reactive, the boundary separating these two fluids, is designed for highest integrity. All welds in these SGs are hence, very critical and require a very high degree of reliability and integrity, especially the thin joints viz. tube to tube sheet joints. Failure of these joints will cause cascading effect. Final integrity of weld joints is required to be ascertained by performing helium leak testing. The paper describes the methodology developed and implementation for leak testing

  16. New procedure for testing HEPA filters: a method omitting testing at 20% rated flow

    International Nuclear Information System (INIS)

    A theoretical analysis of HEPA penetration vs. pinholes in filter media is presented. A calculation formula which shows the tendency of such changes was derived and has been demonstrated experimentally. By this formula equivalent penetration was calculated. Nomograms were used to facilitate the work. As a result, testing needs to be performed only at rated flow

  17. 40 CFR 53.62 - Test procedure: Full wind tunnel test.

    Science.gov (United States)

    2010-07-01

    ... samplers. (iv) Precision calculation. (A) Calculate the coefficient of variation of the mass concentration...) Sampling effectiveness precision. (A) Calculate and record the coefficient of variation for the replicate... percent of the value specified in table F-2 of this subpart, and the variation at any test point in...

  18. 40 CFR 53.63 - Test procedure: Wind tunnel inlet aspiration test.

    Science.gov (United States)

    2010-07-01

    ... the sampler inlet opening centered in the sampling zone. To meet the maximum blockage limit of § 53.62(c)(1) or for convenience, part of the test sampler may be positioned external to the wind tunnel... = reference method sampler volumetric flow rate; and t = sampling time. (iii) Remove the reference...

  19. 75 FR 59105 - Procedures for Transportation Workplace Drug and Alcohol Testing Programs: Federal Drug Testing...

    Science.gov (United States)

    2010-09-27

    ... Federal CCF.'' [74 FR 59196] Because many of the commenters were transportation industry employers, C/TPAs... Step 5A on Copy 1 of the CCF, the new drug analytes MDMA, MDA, and MDEA are added, as are `` 9-THCA... a confirmed positive drug test for MDMA, MDA, or MDEA, as appropriate, to the MRO, the...

  20. Gamma-ray spectrometry combined with acceptable knowledge (GSAK). A technique for characterization of certain remote-handled transuranic (RH-TRU) wastes. Part 2. Testing and results

    International Nuclear Information System (INIS)

    Gamma-ray spectrometry combined with acceptable knowledge (GSAK) is a technique for the characterization of certain remote-handled transuranic (RH-TRU) wastes. GSAK uses gamma-ray spectrometry to quantify a portion of the fission product inventory of RH-TRU wastes. These fission product results are then coupled with calculated inventories derived from acceptable process knowledge to characterize the radionuclide content of the assayed wastes. GSAK has been evaluated and tested through several test exercises. These tests and their results are described; while the former paper in this issue presents the methodology, equipment and techniques. (author)