By measuring organ doses with TLD at an Alderson-Rando-phantom it was tested to evaluate the somatic risk for mammographies. Similarly as the genetic risk can be expressed by the genetic significant dose, the somatic risk can be expressed by the somatic significant dose index. The measurement results of the mammography show a large deviation of almost one power of ten with dependence of the chosen technique. In conventional xeroradiography and recording technique with foilless film the somatic radiation risk is the highest. By using amplifying foils in combination with high-sensitive films it is relatively small. Unfortunately the magnitude of the somatic risk in the different mammography techniques is in the ratio reciprocal to the corresponding image quality. At the time being the most favourable technique is mammography with a soft rastering method, although the xeromammography with additional high aluminium-filtering is a further alternative with respect to ...