WorldWideScience

Sample records for calibration transfer procedures

  1. Validation of a densimeter calibration procedure for a secondary calibration laboratory

    International Nuclear Information System (INIS)

    Alpizar Herrera, Juan Carlos

    2014-01-01

    A survey was conducted to quantify the need for calibration of a density measurement instrument at the research units at the Sede Rodrigo Facio of the Universidad de Costa Rica. A calibration procedure was documented for the instrument that presented the highest demand in the survey by the calibration service. A study of INTE-ISO/IEC 17025: 2005 and specifically in section 5.4 of this standard was done, to document the procedure for calibrating densimeters. Densimeter calibration procedures and standards were sought from different national and international sources. The method of hydrostatic weighing or Cuckow method was the basis of the defined procedure. Documenting the calibration procedure and creating other documents was performed for data acquisition log, intermediate calculation log and calibration certificate copy. A veracity test was performed using as reference laboratory a laboratory of calibration secondary national as part of the validation process of the documented procedure. The results of the E_n statistic of 0.41; 0.34 and 0.46 for the calibration points 90%, 50% and 10% were obtained for the densimeter scale respectively. A reproducibility analysis of the method was performed with satisfactory results. Different suppliers were contacted to estimate the economic costs of the equipment and materials, needed to develop the documented method of densimeter calibration. The acquisition of an analytical balance was recommended, instead of a precision scale, in order to improve the results obtained with the documented method [es

  2. Visible spectroscopy calibration transfer model in determining pH of Sala mangoes

    International Nuclear Information System (INIS)

    Yahaya, O.K.M.; MatJafri, M.Z.; Aziz, A.A.; Omar, A.F.

    2015-01-01

    The purpose of this study is to compare the efficiency of calibration transfer procedures between three spectrometers involving two Ocean Optics Inc. spectrometers, namely, QE65000 and Jaz, and also, ASD FieldSpec 3 in measuring the pH of Sala mango by visible reflectance spectroscopy. This study evaluates the ability of these spectrometers in measuring the pH of Sala mango by applying similar calibration algorithms through direct calibration transfer. This visible reflectance spectroscopy technique defines a spectrometer as a master instrument and another spectrometer as a slave. The multiple linear regression (MLR) of calibration model generated using the QE65000 spectrometer is transferred to the Jaz spectrometer and vice versa for Set 1. The same technique is applied for Set 2 with QE65000 spectrometer is transferred to the FieldSpec3 spectrometer and vice versa. For Set 1, the result showed that the QE65000 spectrometer established a calibration model with higher accuracy than that of the Jaz spectrometer. In addition, the calibration model developed on Jaz spectrometer successfully predicted the pH of Sala mango, which was measured using QE65000 spectrometer, with a root means square error of prediction RMSEP = 0.092 pH and coefficients of determination R 2  = 0.892. Moreover, the best prediction result is obtained for Set 2 when the calibration model developed on QE65000 spectrometer is successfully transferred to FieldSpec 3 with R 2  = 0.839 and RMSEP = 0.16 pH

  3. Optimization of procedure for calibration with radiometer/photometer

    International Nuclear Information System (INIS)

    Detilly, Isabelle

    2009-01-01

    A test procedure for the radiometer/photometer calibrations mark International Light at the Laboratorio de Fotometria y Tecnologia Laser (LAFTA) de la Escuela de Ingenieria Electrica de la Universidad de Costa Rica is established. Two photometric banks are used as experimental set and two calibrations were performed of the International Light. A basic procedure established in the laboratory, is used for calibration from measurements of illuminance and luminous intensity. Some dependent variations of photometric banks used in the calibration process, the programming of the radiometer/photometer and the applied methodology showed the results. The procedure for calibration with radiometer/photometer can be improved by optimizing the programming process of the measurement instrument and possible errors can be minimized by using the recommended procedure. (author) [es

  4. The transfer voltage standard for calibration outside of a laboratory

    Directory of Open Access Journals (Sweden)

    Urekar Marjan

    2017-01-01

    Full Text Available The transfer voltage standard is designed for transferring the analog voltage from a calibrator to the process control workstation for multi-electrode electrolysis process in a plating plant. Transfer voltage standard is based on polypropylene capacitors and operational amplifiers with tera-ohm range input resistance needed for capacitor self-discharging effect cancellation. Dielectric absorption effect is described. An instrument for comparison of reference and control voltages is devised, based on precise window comparator. Detailed description of the main task is given, including constraints, theoretical and practical solutions. Procedure for usage of the standard outside of a laboratory conditions is explained. Comparison of expected and realized standard characteristics is given. [Project of the Serbian Ministry of Education, Science and Technological Development, Grant no. TR-32019

  5. Traceable calibration and demonstration of a portable dynamic force transfer standard

    Science.gov (United States)

    Vlajic, Nicholas; Chijioke, Ako

    2017-08-01

    In general, the dynamic sensitivity of a force transducer depends upon the mechanical system in which it is used. This dependence serves as motivation to develop a dynamic force transfer standard, which can be used to calibrate an application transducer in situ. In this work, we SI-traceably calibrate a hand-held force transducer, namely an impact hammer, by using a mass suspended from a thin line which is cut to produce a known dynamic force in the form of a step function. We show that this instrument is a promising candidate as a transfer standard, since its dynamic response has small variance between different users. This calibrated transfer standard is then used to calibrate a secondary force transducer in an example application setting. The combined standard uncertainty (k  =  2) in the calibration of the transfer standard was determined to be 2.1% or less, up to a bandwidth of 5 kHz. The combined standard uncertainty (k  =  2) in the performed transfer calibration was less than 4%, up to 3 kHz. An advantage of the transfer calibration framework presented here, is that the transfer standard can be used to transfer SI-traceable calibrations without the use of any SI-traceable voltage metrology instrumentation.

  6. Improvement of gamma calibration procedures with commercial management software

    International Nuclear Information System (INIS)

    Lucena, Rodrigo F.; Potiens, Maria da Penha A.; Santos, Gelson P.; Vivolo, Vitor

    2007-01-01

    In this work, the gamma calibration procedure of the Instruments Calibration Laboratory (LCI) of the IPEN-CNEN-SP was improved with the use of the commercial management software Autolab TM from Automa Company. That software was adapted for our specific use in the calibration procedures. The evaluation of the uncertainties in gamma calibration protocol was improved by the LCI staff and yet the all worksheets and final calibration report lay-out was developed in commercial software like Excell TM and Word TM from Microsft TM . (author)

  7. Procedures for calibration of brachytherapy sources

    International Nuclear Information System (INIS)

    Alfonso Laguardia, R.; Alonso Samper, J.L.; Morales Lopez, J.L.; Saez Nunez, D.G.

    1997-01-01

    Brachytherapy source strength verification is a responsibility of the user of these source, in fact of the Medical Physicists in charge of this issue in a Radiotherapy Service. The calibration procedures in the users conditions are shown. Specifics methods for source strength determination are recommended, both for High Dose Rate (HDR) sources with Remote Afterloading equipment and for Low Dose Rate sources. The The results of the calibration of HDR Remote After loaders are indicated

  8. Evaluation of multivariate calibration models transferred between spectroscopic instruments

    DEFF Research Database (Denmark)

    Eskildsen, Carl Emil Aae; Hansen, Per W.; Skov, Thomas

    2016-01-01

    In a setting where multiple spectroscopic instruments are used for the same measurements it may be convenient to develop the calibration model on a single instrument and then transfer this model to the other instruments. In the ideal scenario, all instruments provide the same predictions for the ......In a setting where multiple spectroscopic instruments are used for the same measurements it may be convenient to develop the calibration model on a single instrument and then transfer this model to the other instruments. In the ideal scenario, all instruments provide the same predictions...... for the same samples using the transferred model. However, sometimes the success of a model transfer is evaluated by comparing the transferred model predictions with the reference values. This is not optimal, as uncertainties in the reference method will impact the evaluation. This paper proposes a new method...... for calibration model transfer evaluation. The new method is based on comparing predictions from different instruments, rather than comparing predictions and reference values. A total of 75 flour samples were available for the study. All samples were measured on ten near infrared (NIR) instruments from two...

  9. Ultra-portable field transfer radiometer for vicarious calibration of earth imaging sensors

    Science.gov (United States)

    Thome, Kurtis; Wenny, Brian; Anderson, Nikolaus; McCorkel, Joel; Czapla-Myers, Jeffrey; Biggar, Stuart

    2018-06-01

    A small portable transfer radiometer has been developed as part of an effort to ensure the quality of upwelling radiance from test sites used for vicarious calibration in the solar reflective. The test sites are used to predict top-of-atmosphere reflectance relying on ground-based measurements of the atmosphere and surface. The portable transfer radiometer is designed for one-person operation for on-site field calibration of instrumentation used to determine ground-leaving radiance. The current work describes the detector- and source-based radiometric calibration of the transfer radiometer highlighting the expected accuracy and SI-traceability. The results indicate differences between the detector-based and source-based results greater than the combined uncertainties of the approaches. Results from recent field deployments of the transfer radiometer using a solar radiation based calibration agree with the source-based laboratory calibration within the combined uncertainties of the methods. The detector-based results show a significant difference to the solar-based calibration. The source-based calibration is used as the basis for a radiance-based calibration of the Landsat-8 Operational Land Imager that agrees with the OLI calibration to within the uncertainties of the methods.

  10. Instrument Calibration and Certification Procedure

    Energy Technology Data Exchange (ETDEWEB)

    Davis, R. Wesley [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2016-05-31

    The Amptec 640SL-2 is a 4-wire Kelvin failsafe resistance meter, designed to reliably use very low-test currents for its resistance measurements. The 640SL-1 is a 2-wire version, designed to support customers using the Reynolds Industries type 311 connector. For both versions, a passive (analog) dual function DC Milliameter/Voltmeter allows the user to verify the actual 640SL output current level and the open circuit voltage on the test leads. This procedure includes tests of essential performance parameters. Any malfunction noticed during calibration, whether specifically tested for or not, shall be corrected before calibration continues or is completed.

  11. The simple procedure for the fluxgate magnetometers calibration

    Science.gov (United States)

    Marusenkov, Andriy

    2014-05-01

    The fluxgate magnetometers are widely used in geophysics investigations including the geomagnetic field monitoring at the global network of geomagnetic observatories as well as for electromagnetic sounding of the Earth's crust conductivity. For solving these tasks the magnetometers have to be calibrated with an appropriate level of accuracy. As a particular case, the ways to satisfy the recent requirements to the scaling and orientation errors of 1-second INTERNAGNET magnetometers are considered in the work. The goal of the present study was to choose a simple and reliable calibration method for estimation of scale factors and angular errors of the three-axis magnetometers in the field. There are a large number of the scalar calibration methods, which use a free rotation of the sensor in the calibration field followed by complicated data processing procedures for numerical solution of the high-order equations set. The chosen approach also exploits the Earth's magnetic field as a calibrating signal, but, in contrast to other methods, the sensor has to be oriented in some particular positions in respect to the total field vector, instead of the sensor free rotation. This allows to use very simple and straightforward linear computation formulas and, as a result, to achieve more reliable estimations of the calibrated parameters. The estimation of the scale factors is performed by the sequential aligning of each component of the sensor in two positions: parallel and anti-parallel to the Earth's magnetic field vector. The estimation of non-orthogonality angles between each pair of components is performed after sequential aligning of the components at the angles +/- 45 and +/- 135 degrees of arc in respect to the total field vector. Due to such four positions approach the estimations of the non-orthogonality angles are invariant to the zero offsets and non-linearity of transfer functions of the components. The experimental justifying of the proposed method by means of the

  12. Statistical design of mass spectrometry calibration procedures

    International Nuclear Information System (INIS)

    Bayne, C.K.

    1996-11-01

    The main objective of this task was to agree on calibration procedures to estimate the system parameters (i.e., dead-time correction, ion-counting conversion efficiency, and detector efficiency factors) for SAL's new Finnigan MAT-262 mass spectrometer. SAL will use this mass spectrometer in a clean-laboratory which was opened in December 1995 to measure uranium and plutonium isotopes on environmental samples. The Finnigan MAT-262 mass spectrometer has a multi-detector system with seven Faraday cup detectors and one ion- counter for the measurement of very small signals (e.g. 10 -17 Ampere range). ORNL has made preliminary estimates of the system parameters based on SAL's experimental data measured in late 1994 when the Finnigan instrument was relatively new. SAL generated additional data in 1995 to verify the calibration procedures for estimating the dead-time correction factor, the ion-counting conversion factor and the Faraday cup detector efficiency factors. The system parameters estimated on the present data will have to be reestablished when the Finnigan MAT-262 is moved-to the new clean- laboratory. Different methods will be used to analyzed environmental samples than the current measurement methods being used. For example, the environmental samples will be electroplated on a single filament rather than using the current two filament system. An outline of the calibration standard operating procedure (SOP) is included

  13. A single model procedure for tank calibration function estimation

    International Nuclear Information System (INIS)

    York, J.C.; Liebetrau, A.M.

    1995-01-01

    Reliable tank calibrations are a vital component of any measurement control and accountability program for bulk materials in a nuclear reprocessing facility. Tank volume calibration functions used in nuclear materials safeguards and accountability programs are typically constructed from several segments, each of which is estimated independently. Ideally, the segments correspond to structural features in the tank. In this paper the authors use an extension of the Thomas-Liebetrau model to estimate the entire calibration function in a single step. This procedure automatically takes significant run-to-run differences into account and yields an estimate of the entire calibration function in one operation. As with other procedures, the first step is to define suitable calibration segments. Next, a polynomial of low degree is specified for each segment. In contrast with the conventional practice of constructing a separate model for each segment, this information is used to set up the design matrix for a single model that encompasses all of the calibration data. Estimation of the model parameters is then done using conventional statistical methods. The method described here has several advantages over traditional methods. First, modeled run-to-run differences can be taken into account automatically at the estimation step. Second, no interpolation is required between successive segments. Third, variance estimates are based on all the data, rather than that from a single segment, with the result that discontinuities in confidence intervals at segment boundaries are eliminated. Fourth, the restrictive assumption of the Thomas-Liebetrau method, that the measured volumes be the same for all runs, is not required. Finally, the proposed methods are readily implemented using standard statistical procedures and widely-used software packages

  14. CALIBRATION PROCEDURES ON OBLIQUE CAMERA SETUPS

    Directory of Open Access Journals (Sweden)

    G. Kemper

    2016-06-01

    Full Text Available Beside the creation of virtual animated 3D City models, analysis for homeland security and city planning, the accurately determination of geometric features out of oblique imagery is an important task today. Due to the huge number of single images the reduction of control points force to make use of direct referencing devices. This causes a precise camera-calibration and additional adjustment procedures. This paper aims to show the workflow of the various calibration steps and will present examples of the calibration flight with the final 3D City model. In difference to most other software, the oblique cameras are used not as co-registered sensors in relation to the nadir one, all camera images enter the AT process as single pre-oriented data. This enables a better post calibration in order to detect variations in the single camera calibration and other mechanical effects. The shown sensor (Oblique Imager is based o 5 Phase One cameras were the nadir one has 80 MPIX equipped with a 50 mm lens while the oblique ones capture images with 50 MPix using 80 mm lenses. The cameras are mounted robust inside a housing to protect this against physical and thermal deformations. The sensor head hosts also an IMU which is connected to a POS AV GNSS Receiver. The sensor is stabilized by a gyro-mount which creates floating Antenna –IMU lever arms. They had to be registered together with the Raw GNSS-IMU Data. The camera calibration procedure was performed based on a special calibration flight with 351 shoots of all 5 cameras and registered the GPS/IMU data. This specific mission was designed in two different altitudes with additional cross lines on each flying heights. The five images from each exposure positions have no overlaps but in the block there are many overlaps resulting in up to 200 measurements per points. On each photo there were in average 110 well distributed measured points which is a satisfying number for the camera calibration. In a first

  15. Multi-spectrometer calibration transfer based on independent component analysis.

    Science.gov (United States)

    Liu, Yan; Xu, Hao; Xia, Zhenzhen; Gong, Zhiyong

    2018-02-26

    Calibration transfer is indispensable for practical applications of near infrared (NIR) spectroscopy due to the need for precise and consistent measurements across different spectrometers. In this work, a method for multi-spectrometer calibration transfer is described based on independent component analysis (ICA). A spectral matrix is first obtained by aligning the spectra measured on different spectrometers. Then, by using independent component analysis, the aligned spectral matrix is decomposed into the mixing matrix and the independent components of different spectrometers. These differing measurements between spectrometers can then be standardized by correcting the coefficients within the independent components. Two NIR datasets of corn and edible oil samples measured with three and four spectrometers, respectively, were used to test the reliability of this method. The results of both datasets reveal that spectra measurements across different spectrometers can be transferred simultaneously and that the partial least squares (PLS) models built with the measurements on one spectrometer can predict that the spectra can be transferred correctly on another.

  16. EPMT: a portable transfer standard for telemetry system pressure-transducer calibration

    International Nuclear Information System (INIS)

    Hasbrouck, R.T.

    1977-01-01

    The LLL developed electronic pressure meter (EPMT) is a portable static-pressure calibration instrument for use with the LLL telemetry transducer system at the Nevada Test Site (NTS). It is significantly more accurate and rugged than the bourdon-tube pressure gauge it replaces, and can be incorporated into a field-use, semi-automatic, pressure calibration system. The process by which a transducer is selected for EPMT use from the inventory of field-service-certified transducers and subjected to an extensive preconditioning and calibration procedure is described. By combining this unusual calibration procedure with a unique, statistically based data-reduction routine, the total uncertainty of the measuring process at each calibration point can be determined with high accuracy

  17. Calibration procedure for Slocum glider deployed optical instruments.

    Science.gov (United States)

    Cetinić, Ivona; Toro-Farmer, Gerardo; Ragan, Matthew; Oberg, Carl; Jones, Burton H

    2009-08-31

    Recent developments in the field of the autonomous underwater vehicles allow the wide usage of these platforms as part of scientific experiments, monitoring campaigns and more. The vehicles are often equipped with sensors measuring temperature, conductivity, chlorophyll a fluorescence (Chl a), colored dissolved organic matter (CDOM) fluorescence, phycoerithrin (PE) fluorescence and spectral volume scattering function at 117 degrees, providing users with high resolution, real time data. However, calibration of these instruments can be problematic. Most in situ calibrations are performed by deploying complementary instrument packages or water samplers in the proximity of the glider. Laboratory calibrations of the mounted sensors are difficult due to the placement of the instruments within the body of the vehicle. For the laboratory calibrations of the Slocum glider instruments we developed a small calibration chamber where we can perform precise calibrations of the optical instruments aboard our glider, as well as sensors from other deployment platforms. These procedures enable us to obtain pre- and post-deployment calibrations for optical fluorescence instruments, which may differ due to the biofouling and other physical damage that can occur during long-term glider deployments. We found that biofouling caused significant changes in the calibration scaling factors of fluorescent sensors, suggesting the need for consistent and repetitive calibrations for gliders as proposed in this paper.

  18. Mathematical calibration procedure of a capacitive sensor-based indexed metrology platform

    International Nuclear Information System (INIS)

    Brau-Avila, A; Valenzuela-Galvan, M; Herrera-Jimenez, V M; Santolaria, J; Aguilar, J J; Acero, R

    2017-01-01

    The demand for faster and more reliable measuring tasks for the control and quality assurance of modern production systems has created new challenges for the field of coordinate metrology. Thus, the search for new solutions in coordinate metrology systems and the need for the development of existing ones still persists. One example of such a system is the portable coordinate measuring machine (PCMM), the use of which in industry has considerably increased in recent years, mostly due to its flexibility for accomplishing in-line measuring tasks as well as its reduced cost and operational advantages compared to traditional coordinate measuring machines. Nevertheless, PCMMs have a significant drawback derived from the techniques applied in the verification and optimization procedures of their kinematic parameters. These techniques are based on the capture of data with the measuring instrument from a calibrated gauge object, fixed successively in various positions so that most of the instrument measuring volume is covered, which results in time-consuming, tedious and expensive verification and optimization procedures. In this work the mathematical calibration procedure of a capacitive sensor-based indexed metrology platform (IMP) is presented. This calibration procedure is based on the readings and geometric features of six capacitive sensors and their targets with nanometer resolution. The final goal of the IMP calibration procedure is to optimize the geometric features of the capacitive sensors and their targets in order to use the optimized data in the verification procedures of PCMMs. (paper)

  19. Mathematical calibration procedure of a capacitive sensor-based indexed metrology platform

    Science.gov (United States)

    Brau-Avila, A.; Santolaria, J.; Acero, R.; Valenzuela-Galvan, M.; Herrera-Jimenez, V. M.; Aguilar, J. J.

    2017-03-01

    The demand for faster and more reliable measuring tasks for the control and quality assurance of modern production systems has created new challenges for the field of coordinate metrology. Thus, the search for new solutions in coordinate metrology systems and the need for the development of existing ones still persists. One example of such a system is the portable coordinate measuring machine (PCMM), the use of which in industry has considerably increased in recent years, mostly due to its flexibility for accomplishing in-line measuring tasks as well as its reduced cost and operational advantages compared to traditional coordinate measuring machines. Nevertheless, PCMMs have a significant drawback derived from the techniques applied in the verification and optimization procedures of their kinematic parameters. These techniques are based on the capture of data with the measuring instrument from a calibrated gauge object, fixed successively in various positions so that most of the instrument measuring volume is covered, which results in time-consuming, tedious and expensive verification and optimization procedures. In this work the mathematical calibration procedure of a capacitive sensor-based indexed metrology platform (IMP) is presented. This calibration procedure is based on the readings and geometric features of six capacitive sensors and their targets with nanometer resolution. The final goal of the IMP calibration procedure is to optimize the geometric features of the capacitive sensors and their targets in order to use the optimized data in the verification procedures of PCMMs.

  20. Investigation of the transferability of hydrological models and a method to improve model calibration

    Directory of Open Access Journals (Sweden)

    G. Hartmann

    2005-01-01

    Full Text Available In order to find a model parameterization such that the hydrological model performs well even under different conditions, appropriate model performance measures have to be determined. A common performance measure is the Nash Sutcliffe efficiency. Usually it is calculated comparing observed and modelled daily values. In this paper a modified version is suggested in order to calibrate a model on different time scales simultaneously (days up to years. A spatially distributed hydrological model based on HBV concept was used. The modelling was applied on the Upper Neckar catchment, a mesoscale river in south western Germany with a basin size of about 4000 km2. The observation period 1961-1990 was divided into four different climatic periods, referred to as "warm", "cold", "wet" and "dry". These sub periods were used to assess the transferability of the model calibration and of the measure of performance. In a first step, the hydrological model was calibrated on a certain period and afterwards applied on the same period. Then, a validation was performed on the climatologically opposite period than the calibration, e.g. the model calibrated on the cold period was applied on the warm period. Optimal parameter sets were identified by an automatic calibration procedure based on Simulated Annealing. The results show, that calibrating a hydrological model that is supposed to handle short as well as long term signals becomes an important task. Especially the objective function has to be chosen very carefully.

  1. Calibration procedure and wavelength correction for neutron depolarization experiments

    International Nuclear Information System (INIS)

    Roest, W.; Rekveldt, M.T.

    1992-01-01

    The neutron polarimeter, for which an extended calibration procedure is described here, enables one to investigate magnetic properties of materials. Such an investigation is carried out by offering a polarized neutron beam in the x-, y- and z-direction successively and, after transmission through the sample, by analysing the polarization in all three directions. The result is a 3x3 depolarization matrix. After the polarizer, the neutron beam has a polarization along the z-direction. Two coil systems creating a magnetic field in the yz-plane perpendicular to the beam direction provide the possibility to direct the polarization in the x-, y- and z-direction by means of Larmor precession of the polarization in these fields. New research areas, where small depolarization effects together with considerable polarization rotation are measured, have caused a need for more accuracy in, and better knowledge of the calibration of the polarimeter. The calibration procedure use up to now and the improvements made on it are described. (orig.)

  2. Optimized procedure for calibration and verification multileaf collimator from Elekta Synergy accelerator

    International Nuclear Information System (INIS)

    Castel Millan, A.; Perellezo Mazon, A.; Fernandez Ibiza, J.; Arnalte Olloquequi, M.; Armengol Martinez, S.; Rodriguez Rey, A.; Guedea Edo, F.

    2011-01-01

    The objective of this work is to design an optimized procedure for calibration and verification of a multileaf collimator used so as to allow the EPID and the image plate in a complementary way, using different processing systems. With this procedure we have two equivalent alternative as the same parameters obtained for the calibration of multileaf Elekta Synergy accelerator.

  3. Computational integration of the phases and procedures of calibration processes for radioprotection

    International Nuclear Information System (INIS)

    Santos, Gleice R. dos; Thiago, Bibiana dos S.; Rocha, Felicia D.G.; Santos, Gelson P. dos; Potiens, Maria da Penha A.; Vivolo, Vitor

    2011-01-01

    This work proceed the computational integration of the processes phases by using only a single computational software, from the entrance of the instrument at the Instrument Calibration Laboratory (LCI-IPEN) to the conclusion of calibration procedures. So, the initial information such as trade mark, model, manufacturer, owner, and the calibration records are digitized once until the calibration certificate emission

  4. Calibration and verification of surface contamination meters --- Procedures and techniques

    International Nuclear Information System (INIS)

    Schuler, C; Butterweck, G.; Wernli, C.; Bochud, F.; Valley, J.-F.

    2007-03-01

    A standardised measurement procedure for surface contamination meters (SCM) is presented. The procedure aims at rendering surface contamination measurements to be simply and safely interpretable. Essential for the approach is the introduction and common use of the radionuclide specific quantity 'guideline value' specified in the Swiss Radiation Protection Ordinance as unit for the measurement of surface activity. The according radionuclide specific 'guideline value count rate' can be summarized as verification reference value for a group of radionuclides ('basis guideline value count rate'). The concept can be generalized for SCM of the same type or for SCM of different types using he same principle of detection. A SCM multi source calibration technique is applied for the determination of the instrument efficiency. Four different electron radiation energy regions, four different photon radiation energy regions and an alpha radiation energy region are represented by a set of calibration sources built according to ISO standard 8769-2. A guideline value count rate representing the activity per unit area of a surface contamination of one guideline value can be calculated for any radionuclide using instrument efficiency, radionuclide decay data, contamination source efficiency, guideline value averaging area (100 cm 2 ), and radionuclide specific guideline value. n this way, instrument responses for the evaluation of surface contaminations are obtained for radionuclides without available calibration sources as well as for short-Iived radionuclides, for which the continuous replacement of certified calibration sources can lead to unreasonable costs. SCM verification is based on surface emission rates of reference sources with an active area of 100 cm 2 . The verification for a given list of radionuclides is based on the radionuclide specific quantity guideline value count rate. Guideline value count rates for groups of radionuclides can be represented within the maximum

  5. Brookhaven National Laboratory meteorological services instrument calibration plan and procedures

    Energy Technology Data Exchange (ETDEWEB)

    Heiser, John [Brookhaven National Lab. (BNL), Upton, NY (United States)

    2013-02-16

    This document describes the Meteorological Services (Met Services) Calibration and Maintenance Schedule and Procedures, The purpose is to establish the frequency and mechanism for the calibration and maintenance of the network of meteorological instrumentation operated by Met Services. The goal is to maintain the network in a manner that will result in accurate, precise and reliable readings from the instrumentation.

  6. Using genetic algorithm and TOPSIS for Xinanjiang model calibration with a single procedure

    Science.gov (United States)

    Cheng, Chun-Tian; Zhao, Ming-Yan; Chau, K. W.; Wu, Xin-Yu

    2006-01-01

    Genetic Algorithm (GA) is globally oriented in searching and thus useful in optimizing multiobjective problems, especially where the objective functions are ill-defined. Conceptual rainfall-runoff models that aim at predicting streamflow from the knowledge of precipitation over a catchment have become a basic tool for flood forecasting. The parameter calibration of a conceptual model usually involves the multiple criteria for judging the performances of observed data. However, it is often difficult to derive all objective functions for the parameter calibration problem of a conceptual model. Thus, a new method to the multiple criteria parameter calibration problem, which combines GA with TOPSIS (technique for order performance by similarity to ideal solution) for Xinanjiang model, is presented. This study is an immediate further development of authors' previous research (Cheng, C.T., Ou, C.P., Chau, K.W., 2002. Combining a fuzzy optimal model with a genetic algorithm to solve multi-objective rainfall-runoff model calibration. Journal of Hydrology, 268, 72-86), whose obvious disadvantages are to split the whole procedure into two parts and to become difficult to integrally grasp the best behaviors of model during the calibration procedure. The current method integrates the two parts of Xinanjiang rainfall-runoff model calibration together, simplifying the procedures of model calibration and validation and easily demonstrated the intrinsic phenomenon of observed data in integrity. Comparison of results with two-step procedure shows that the current methodology gives similar results to the previous method, is also feasible and robust, but simpler and easier to apply in practice.

  7. Advanced Measurement and Simulation Procedure for the Identification of Heat and Mass Transfer Parameters in Dynamic Adsorption Experiments

    Directory of Open Access Journals (Sweden)

    Andreas Velte

    2017-08-01

    Full Text Available Thermally-driven heat pumps can help to mitigate CO2 emissions by enhancing the efficiency of heating systems or by driving cooling systems with waste or solar heat. In order to make the thermally-driven systems more attractive for the end consumer, these systems need a higher power density. A higher power density can be achieved by intensifying the heat and mass transfer processes within the adsorption heat exchanger. For the optimization of this key component, a numerical model of the non-isothermal adsorption dynamics can be applied. The calibration of such a model can be difficult, since heat and mass transfer processes are strongly coupled. We present a measurement and simulation procedure that makes it possible to calibrate the heat transfer part of the numerical model separately from the mass transfer part. Furthermore, it is possible to identify the parts of the model that need to be improved. For this purpose, a modification of the well-known large temperature jump method is developed. The newly-introduced measurements are conducted under an inert N2 atmosphere, and the surface temperature of the sample is measured with an infrared sensor. We show that the procedure is applicable for two completely different types of samples: a loose grains configuration and a fibrous structure that is directly crystallized.

  8. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 1: Procedural overview

    International Nuclear Information System (INIS)

    2007-01-01

    Accurate determinations of volume are a fundamental component of any measurement-based system of control and accountability in a facility that processes or stores nuclear materials in liquid form. Volume determinations are typically made with the aid of a calibration or volume measurement equation that relates the response of the tank's measurement system to some independent measure of tank volume. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation (the inverse of the calibration equation), which relates tank volume to measurement system response. The steps carried out to acquire data for estimating the tank's calibration or volume measurement equation are collectively described as the process of tank calibration. This part of ISO 18213 describes procedures for tank calibration and volume determination for nuclear process tanks equipped with pressure-measurement systems for determining liquid content. Specifically, overall guidance is provided for planning a calibration exercise undertaken to obtain the data required for the measurement equation to estimate a tank's volume. The key steps in the procedure are also presented for subsequently using the estimated volume-measurement equation to determine tank liquid volumes. The procedures presented apply specifically to tanks equipped with bubbler probe systems for measuring liquid content. Moreover, these procedures produce reliable results only for clear (i.e. without suspended solids), homogeneous liquids that are at both thermal and static equilibrium. The paper elaborates on scope, physical principles involved, the calibration model, equipment required, a typical tank calibration procedure, calibration planning and pre-calibration activities, and volume determination. A bibliography is provided

  9. Permanently calibrated interpolating time counter

    International Nuclear Information System (INIS)

    Jachna, Z; Szplet, R; Kwiatkowski, P; Różyc, K

    2015-01-01

    We propose a new architecture of an integrated time interval counter that provides its permanent calibration in the background. Time interval measurement and the calibration procedure are based on the use of a two-stage interpolation method and parallel processing of measurement and calibration data. The parallel processing is achieved by a doubling of two-stage interpolators in measurement channels of the counter, and by an appropriate extension of control logic. Such modification allows the updating of transfer characteristics of interpolators without the need to break a theoretically infinite measurement session. We describe the principle of permanent calibration, its implementation and influence on the quality of the counter. The precision of the presented counter is kept at a constant level (below 20 ps) despite significant changes in the ambient temperature (from −10 to 60 °C), which can cause a sevenfold decrease in the precision of the counter with a traditional calibration procedure. (paper)

  10. Calibration of IR test chambers with the missile defense transfer radiometer

    Science.gov (United States)

    Kaplan, Simon G.; Woods, Solomon I.; Carter, Adriaan C.; Jung, Timothy M.

    2013-05-01

    The Missile Defense Transfer Radiometer (MDXR) is designed to calibrate infrared collimated and flood sources over the fW/cm2 to W/cm2 power range from 3 μm to 28μ m in wavelength. The MDXR operates in three different modes: as a filter radiometer, a Fourier-transform spectrometer (FTS)-based spectroradiometer, and as an absolute cryogenic radiometer (ACR). Since 2010, the MDXR has made measurements of the collimated infrared irradiance at the output port of seven different infrared test chambers at several facilities. We present a selection of results from these calibration efforts compared to signal predictions from the respective chamber models for the three different MDXR calibration modes. We also compare the results to previous measurements made of the same chambers with a legacy transfer radiometer, the NIST BXR. In general, the results are found to agree within their combined uncertainties, with the MDXR having 30 % lower uncertainty and greater spectral coverage.

  11. Transfer of Procedure in Criminal Matters in Romanian Legislation

    Directory of Open Access Journals (Sweden)

    Ion Rusu

    2009-06-01

    Full Text Available Aware of the obligations assumed in fighting crime, Romania translated in itsinternal legislation the European Convention on the transfer of procedure in criminal matters,adopted in Strasbourg on 15 May 1972, ratified through Government’s Ordinance no.77/1999.The transfer of procedure in criminal matters is one of the forms of international judicialcooperation in criminal matter and represents an act on mutual trust in the organizing activityfor crime pursuit at the international level. According to law, the procedure transfer in criminalmatter consists in performing criminal procedure or continuing the procedures initiated by thecompetent Romanian authorities for an action that represents a crime, in accordance with theRomanian law and transferring it to another state. The procedure transfer in criminal matters isaccomplished only if the conditions expressly provisioned by law are fulfilled, respecting thenon bis in idem principle.

  12. High-precision reflectivity measurements: improvements in the calibration procedure

    Science.gov (United States)

    Jupe, Marco; Grossmann, Florian; Starke, Kai; Ristau, Detlev

    2003-05-01

    The development of high quality optical components is heavily depending on precise characterization procedures. The reflectance and transmittance of laser components are the most important parameters for advanced laser applications. In the industrial fabrication of optical coatings, quality management is generally insured by spectral photometric methods according to ISO/DIS 15386 on a medium level of accuracy. Especially for high reflecting mirrors, a severe discrepancy in the determination of the absolute reflectivity can be found for spectral photometric procedures. In the first part of the CHOCLAB project, a method for measuring reflectance and transmittance with an enhanced precision was developed, which is described in ISO/WD 13697. In the second part of the CHOCLAB project, the evaluation and optimization for the presented method is scheduled. Within this framework international Round-Robin experiment is currently in progress. During this Round-Robin experiment, distinct deviations could be observed between the results of high precision measurement facilities of different partners. Based on the extended experiments, the inhomogeneity of the sample reflectivity was identified as one important origin for the deviation. Consequently, this inhomogeneity is also influencing the calibration procedure. Therefore, a method was developed that allows the calibration of the chopper blade using always the same position on the reference mirror. During the investigations, the homogeneity of several samples was characterized by a surface mapping procedure for 1064 nm. The measurement facility was extended to the additional wavelength 532 nm and a similar set-up was assembled at 10.6 μm. The high precision reflectivity procedure at the mentioned wavelengths is demonstrated for exemplary measurements.

  13. QA experience at the University of Wisconsin accredited dosimetry calibration laboratory

    Energy Technology Data Exchange (ETDEWEB)

    DeWard, L.A.; Micka, J.A. [Univ. of Wisconsin, Madison, WI (United States)

    1993-12-31

    The University of Wisconsin Accredited Dosimetry Calibration Laboratory (UW ADCL) employs procedure manuals as part of its Quality Assurance (QA) program. One of these manuals covers the QA procedures and results for all of the UW ADCL measurement equipment. The QA procedures are divided into two main areas: QA for laboratory equipment and QA for external chambers sent for calibration. All internal laboratory equipment is checked and recalibrated on an annual basis, after establishing its consistency on a 6-month basis. QA for external instruments involves checking past calibration history as well as comparing to a range of calibration values for specific instrument models. Generally, the authors find that a chamber will have a variation of less than 0.5 % from previous Co-60 calibration factors, and falls within two standard deviations of previous calibrations. If x-ray calibrations are also performed, the energy response of the chamber is plotted and compared to previous instruments of the same model. These procedures give the authors confidence in the transfer of calibration values from National Institute of Standards and Technology (NIST).

  14. QA experience at the University of Wisconsin accredited dosimetry calibration laboratory

    International Nuclear Information System (INIS)

    DeWard, L.A.; Micka, J.A.

    1993-01-01

    The University of Wisconsin Accredited Dosimetry Calibration Laboratory (UW ADCL) employs procedure manuals as part of its Quality Assurance (QA) program. One of these manuals covers the QA procedures and results for all of the UW ADCL measurement equipment. The QA procedures are divided into two main areas: QA for laboratory equipment and QA for external chambers sent for calibration. All internal laboratory equipment is checked and recalibrated on an annual basis, after establishing its consistency on a 6-month basis. QA for external instruments involves checking past calibration history as well as comparing to a range of calibration values for specific instrument models. Generally, the authors find that a chamber will have a variation of less than 0.5 % from previous Co-60 calibration factors, and falls within two standard deviations of previous calibrations. If x-ray calibrations are also performed, the energy response of the chamber is plotted and compared to previous instruments of the same model. These procedures give the authors confidence in the transfer of calibration values from National Institute of Standards and Technology (NIST)

  15. Calibration and correction procedures for cosmic-ray neutron soil moisture probes located across Australia

    Science.gov (United States)

    Hawdon, Aaron; McJannet, David; Wallace, Jim

    2014-06-01

    The cosmic-ray probe (CRP) provides continuous estimates of soil moisture over an area of ˜30 ha by counting fast neutrons produced from cosmic rays which are predominantly moderated by water molecules in the soil. This paper describes the setup, measurement correction procedures, and field calibration of CRPs at nine locations across Australia with contrasting soil type, climate, and land cover. These probes form the inaugural Australian CRP network, which is known as CosmOz. CRP measurements require neutron count rates to be corrected for effects of atmospheric pressure, water vapor pressure changes, and variations in incoming neutron intensity. We assess the magnitude and importance of these corrections and present standardized approaches for network-wide analysis. In particular, we present a new approach to correct for incoming neutron intensity variations and test its performance against existing procedures used in other studies. Our field calibration results indicate that a generalized calibration function for relating neutron counts to soil moisture is suitable for all soil types, with the possible exception of very sandy soils with low water content. Using multiple calibration data sets, we demonstrate that the generalized calibration function only applies after accounting for persistent sources of hydrogen in the soil profile. Finally, we demonstrate that by following standardized correction procedures and scaling neutron counting rates of all CRPs to a single reference location, differences in calibrations between sites are related to site biomass. This observation provides a means for estimating biomass at a given location or for deriving coefficients for the calibration function in the absence of field calibration data.

  16. Metrological reliability of the calibration procedure in terms of air kerma using the ionization chamber NE2575

    International Nuclear Information System (INIS)

    Guimaraes, Margarete Cristina; Silva, Teogenes Augusto da; Rosado, Paulo H.G.

    2016-01-01

    Metrology laboratories are expected to provide X radiation beams that were established by international standardization organizations to perform calibration and testing of dosimeters. Reliable and traceable standard dosimeters should be used in the calibration procedure. The aim of this work was to study the reliability of the NE 2575 ionization chamber used as standard dosimeter for the air kerma calibration procedure adopted in the CDTN Calibration Laboratory. (author)

  17. Neutron activation procedures used for the calibration of a nitrogen-16 reactor power monitor

    International Nuclear Information System (INIS)

    Jester, W.A.; Daubenspeck, T.

    2000-01-01

    Personnel from the Pennsylvania State University's Breazeale Nuclear Reactor assisted NRC (Nuclear Research Corporation) personnel in calibrating a new MSL/ 16 N Nitrogen-16 Monitor. Neutron flux calibration procedures utilized and the results obtained for the production of a nitrogen-16 source of known activity for a BGO detector calibration are described. (author)

  18. Calibration transfer between electronic nose systems for rapid In situ measurement of pulp and paper industry emissions

    Energy Technology Data Exchange (ETDEWEB)

    Deshmukh, Sharvari [CSIR-National Environmental Engineering and Research Institute, Nagpur (India); Department of Instrumentation and Electronics Engineering, Jadavpur University, Kolkata (India); Kamde, Kalyani [CSIR-National Environmental Engineering and Research Institute, Nagpur (India); Jana, Arun [Center for Development of Advance Computing, Kolkata (India); Korde, Sanjivani [CSIR-National Environmental Engineering and Research Institute, Nagpur (India); Bandyopadhyay, Rajib [Department of Instrumentation and Electronics Engineering, Jadavpur University, Kolkata (India); Sankar, Ravi [Center for Development of Advance Computing, Kolkata (India); Bhattacharyya, Nabarun, E-mail: nabarun.bhattacharya@cdac.in [Center for Development of Advance Computing, Kolkata (India); Pandey, R.A., E-mail: ra_pandey@neeri.res.in [CSIR-National Environmental Engineering and Research Institute, Nagpur (India)

    2014-09-02

    Highlights: • E-nose developed for obnoxious emissions measurement at pulp and paper industrial site. • ANN model developed for prediction of (CH{sub 3}){sub 2}S, (CH{sub 3}){sub 2}S{sub 2}, CH{sub 3}SH and H{sub 2}S concentration. • Calibration transfer methodology developed for transfer between two e-nose instruments. • Box–Behnken design and robust regression used for calibration transfer. • Results show effective transfer of training model from one e-nose system to other. - Abstract: Electronic nose systems when deployed in network mesh can effectively provide a low budget and onsite solution for the industrial obnoxious gaseous measurement. For accurate and identical prediction capability by all the electronic nose systems, a reliable calibration transfer model needs to be implemented in order to overcome the inherent sensor array variability. In this work, robust regression (RR) is used for calibration transfer between two electronic nose systems using a Box–Behnken (BB) design. Out of the two electronic nose systems, one was trained using industrial gas samples by four artificial neural network models, for the measurement of obnoxious odours emitted from pulp and paper industries. The emissions constitute mainly of hydrogen sulphide (H{sub 2}S), methyl mercaptan (MM), dimethyl sulphide (DMS) and dimethyl disulphide (DMDS) in different proportions. A Box–Behnken design consisting of 27 experiment sets based on synthetic gas combinations of H{sub 2}S, MM, DMS and DMDS, were conducted for calibration transfer between two identical electronic nose systems. Identical sensors on both the systems were mapped and the prediction models developed using ANN were then transferred to the second system using BB–RR methodology. The results showed successful transmission of prediction models developed for one system to other system, with the mean absolute error between the actual and predicted concentration of analytes in mg L{sup −1} after calibration

  19. Calibration transfer between electronic nose systems for rapid In situ measurement of pulp and paper industry emissions

    International Nuclear Information System (INIS)

    Deshmukh, Sharvari; Kamde, Kalyani; Jana, Arun; Korde, Sanjivani; Bandyopadhyay, Rajib; Sankar, Ravi; Bhattacharyya, Nabarun; Pandey, R.A.

    2014-01-01

    Highlights: • E-nose developed for obnoxious emissions measurement at pulp and paper industrial site. • ANN model developed for prediction of (CH 3 ) 2 S, (CH 3 ) 2 S 2 , CH 3 SH and H 2 S concentration. • Calibration transfer methodology developed for transfer between two e-nose instruments. • Box–Behnken design and robust regression used for calibration transfer. • Results show effective transfer of training model from one e-nose system to other. - Abstract: Electronic nose systems when deployed in network mesh can effectively provide a low budget and onsite solution for the industrial obnoxious gaseous measurement. For accurate and identical prediction capability by all the electronic nose systems, a reliable calibration transfer model needs to be implemented in order to overcome the inherent sensor array variability. In this work, robust regression (RR) is used for calibration transfer between two electronic nose systems using a Box–Behnken (BB) design. Out of the two electronic nose systems, one was trained using industrial gas samples by four artificial neural network models, for the measurement of obnoxious odours emitted from pulp and paper industries. The emissions constitute mainly of hydrogen sulphide (H 2 S), methyl mercaptan (MM), dimethyl sulphide (DMS) and dimethyl disulphide (DMDS) in different proportions. A Box–Behnken design consisting of 27 experiment sets based on synthetic gas combinations of H 2 S, MM, DMS and DMDS, were conducted for calibration transfer between two identical electronic nose systems. Identical sensors on both the systems were mapped and the prediction models developed using ANN were then transferred to the second system using BB–RR methodology. The results showed successful transmission of prediction models developed for one system to other system, with the mean absolute error between the actual and predicted concentration of analytes in mg L −1 after calibration transfer (on second system) being 0.076, 0

  20. Design, fabrication, and calibration of curved integral coils for measuring transfer function, uniformity, and effective length of LBL ALS [Lawrence Berkeley Laboratory Advanced Light Source] Booster Dipole Magnets

    International Nuclear Information System (INIS)

    Green, M.I.; Nelson, D.; Marks, S.; Gee, B.; Wong, W.; Meneghetti, J.

    1989-03-01

    A matched pair of curved integral coils has been designed, fabricated and calibrated at Lawrence Berkeley Laboratory for measuring Advanced Light Source (ALS) Booster Dipole Magnets. Distinctive fabrication and calibration techniques are described. The use of multifilar magnet wire in fabrication integral search coils is described. Procedures used and results of AC and DC measurements of transfer function, effective length and uniformity of the prototype booster dipole magnet are presented in companion papers. 8 refs

  1. A multivariate calibration procedure for the tensammetric determination of detergents

    NARCIS (Netherlands)

    Bos, M.

    1989-01-01

    A multivariate calibration procedure based on singular value decomposition (SVD) and the Ho-Kashyap algorithm is used for the tensammetric determination of the cationic detergents Hyamine 1622, benzalkonium chloride (BACl), N-cetyl-N,N,N-trimethylammonium bromide (CTABr) and mixtures of CTABr and

  2. Learning a common dictionary for subject-transfer decoding with resting calibration.

    Science.gov (United States)

    Morioka, Hiroshi; Kanemura, Atsunori; Hirayama, Jun-ichiro; Shikauchi, Manabu; Ogawa, Takeshi; Ikeda, Shigeyuki; Kawanabe, Motoaki; Ishii, Shin

    2015-05-01

    Brain signals measured over a series of experiments have inherent variability because of different physical and mental conditions among multiple subjects and sessions. Such variability complicates the analysis of data from multiple subjects and sessions in a consistent way, and degrades the performance of subject-transfer decoding in a brain-machine interface (BMI). To accommodate the variability in brain signals, we propose 1) a method for extracting spatial bases (or a dictionary) shared by multiple subjects, by employing a signal-processing technique of dictionary learning modified to compensate for variations between subjects and sessions, and 2) an approach to subject-transfer decoding that uses the resting-state activity of a previously unseen target subject as calibration data for compensating for variations, eliminating the need for a standard calibration based on task sessions. Applying our methodology to a dataset of electroencephalography (EEG) recordings during a selective visual-spatial attention task from multiple subjects and sessions, where the variability compensation was essential for reducing the redundancy of the dictionary, we found that the extracted common brain activities were reasonable in the light of neuroscience knowledge. The applicability to subject-transfer decoding was confirmed by improved performance over existing decoding methods. These results suggest that analyzing multisubject brain activities on common bases by the proposed method enables information sharing across subjects with low-burden resting calibration, and is effective for practical use of BMI in variable environments. Copyright © 2015 Elsevier Inc. All rights reserved.

  3. Two calibration procedures for a gyroscope-free inertial measurement system based on a double-pendulum apparatus

    Science.gov (United States)

    Cappa, P.; Patanè, F.; Rossi, S.

    2008-05-01

    This paper presents a novel calibration algorithm to be used with a gyro-free inertial measurement unit (GF-IMU) based on the use of linear accelerometers (AC). The analytical approach can be implemented in two calibration procedures. The first procedure (P-I) is articulated in the conduction of a static trial, to compute the sensitivity and the direction of the sensing axis of each AC, followed by a dynamic trial, to determine the AC locations. By contrast, the latter procedure (P-II) consists in the calculation of the previously indicated calibration parameters by means of a dynamic trial only. The feasibility of the two calibration procedures has been investigated by testing two GF-IMUs, equipped with ten and six bi-axial linear ACs, with an ad hoc instrumented double-pendulum apparatus. P-I and P-II were compared to a calibration procedure used as a reference (P-REF), which incorporates the AC positions measured with an optoelectronic system. The experimental results we present in this paper demonstrate that (i) P-I is able to determine the calibration parameters of the AC array with a higher accuracy than P-II; (ii) consequently, the errors associated with translational (a0 - g) and rotational (\\dot{\\bm \\omega }) acceleration components for the two GF-IMUs are significantly greater using P-II than P-I and (iii) the errors in (a0 - g) and \\dot{\\bm \\omega } obtained with P-I are comparable with the ones obtainable by using P-REF. Thus, the proposed novel algorithm used in P-I, in conjunction with the double-pendulum apparatus, can be globally considered a viable tool in GF-IMU calibration.

  4. The calibration procedures in the Studsvik standardized personnel dosimetry system

    International Nuclear Information System (INIS)

    Widell, C.O.

    1978-01-01

    Every large nuclear installation in Sweden reads its own personnel TLDs. In order to supervise this decentralized reading of dose meters, the TLD readers are connected by telephone lines to a central computer for dose registration. This computer is used both for registering the personnel doses and for checking the TLD readers. This checking is performed by the use of pre-irradiated calibration dose meters which are always used when a batch of personnel dose meters are read. The pre-irradiated dose meters are either irradiated using 137 Cs to various doses up to 100mSv(10000mrem) or using a 90 Sr source in a reference dose irradiator to a dose equal to 3mSv(300mrem) from a 137 Cs source. The results from the reading of the pre-irradiated dose meters are processed by the computer and a calibration factor is calculated. The calibration factor is automatically used to calculate the doses to the personnel TLD's. However, if the calibration factor deviates by more than 10% from the previously used factor, this fact is shown to the operator - who then has to decide what calibration factor is going to be used. This calibration and supervisory procedure together with the safety interlocks in the TLD readers has resulted in a very reliable and accurate dosimetry system. (author)

  5. Initial tank calibration at NUCEF critical facility. 1. Measurement procedure and its result

    International Nuclear Information System (INIS)

    Yanagisawa, Hiroshi; Mineo, Hideaki; Tonoike, Kotaro; Takeshita, Isao; Hoshi, Katsuya; Hagiwara, Hiroyuki.

    1994-07-01

    Initial tank calibrations were carried out prior to hot operation of critical facilities in NUCEF: Nuclear Fuel Cycle Safety Engineering Research Facility, for the purpose of the nuclear material accountancy and control for the facility. Raw calibration data were collected from single run per one tank by measuring differential pressure with dip-tube systems, weight of calibration liquid (demineralized water) poured into the tank, temperature in the tank and so on, without operation of tank ventilation system. Volume and level data were obtained by applying density and buoyancy corrections to the raw data. As a result, the evaluated measurement errors of volume and level were small enough, e.g. within 0.2 lit. and 1.0 mm, respectively, for Pu accountancy tanks. This paper summarizes the above-mentioned measurement procedures, collected data, data correction procedures and evaluated measurement errors. (author)

  6. Effects of pressurization procedures on calibration results for precise pressure transducers

    International Nuclear Information System (INIS)

    Kajikawa, Hiroaki; Kobata, Tokihiko

    2010-01-01

    The output of electromechanical pressure gauges depends on not only the currently applied pressure, but also the pressurization history. Thus, the calibration results of gauges are affected by the pressurization procedure. In this paper, among several important factors influencing the results, we report the effects of the interval between the calibration cycles and the effects of the preliminary pressurizations. In order to quantitatively evaluate these effects, we developed a fully automated system that uses a pressure balance to calibrate pressure gauges. Subsequently, gauges containing quartz Bourdon-type pressure transducers were calibrated in a stepwise manner for pressures between 10 MPa and 100 MPa. The typical standard deviation of the data over three cycles was reduced to a few parts per million (ppm). The interval between the calibration cycles, which ranges from zero to more than 12 h, exerts a strong influence on the results in the process of increasing the pressure, where at 10 MPa the maximum difference between the results was approximately 40 ppm. The preliminary pressurization immediately before the calibration cycle reduces the effects of the interval on the results in certain cases. However, in turn, the influence of the waiting time between the preliminary pressurization and the main calibration cycle becomes strong. In the present paper, we outline several possible measures for obtaining calibration results with high reproducibility

  7. Validation of the in-flight calibration procedures for the MICROSCOPE space mission

    Science.gov (United States)

    Hardy, Émilie; Levy, Agnès; Rodrigues, Manuel; Touboul, Pierre; Métris, Gilles

    2013-11-01

    The MICROSCOPE space mission aims to test the Equivalence Principle with an accuracy of 10-15. The drag-free micro-satellite will orbit around the Earth and embark a differential electrostatic accelerometer including two cylindrical test masses submitted to the same gravitational field and made of different materials. The experience consists in testing the equality of the electrostatic acceleration applied to the masses to maintain them relatively motionless. The accuracy of the measurements exploited for the test of the Equivalence Principle is limited by our a priori knowledge of several physical parameters of the instrument. These parameters are partially estimated on-ground, but with an insufficient accuracy, and an in-orbit calibration is therefore required to correct the measurements. The calibration procedures have been defined and their analytical performances have been evaluated. In addition, a simulator software including the dynamics model of the instrument, the satellite drag-free system and the perturbing environment has been developed to numerically validate the analytical results. After an overall presentation of the MICROSCOPE mission, this paper will describe the calibration procedures and focus on the simulator. Such an in-flight calibration is mandatory for similar space missions taking advantage of a drag-free system.

  8. Providing radiometric traceability for the calibration home base of DLR by PTB

    Energy Technology Data Exchange (ETDEWEB)

    Taubert, D. R.; Hollandt, J.; Sperfeld, P.; Pape, S.; Hoepe, A.; Hauer, K.-O. [Physikalisch-Technische Bundesanstalt, Braunschweig und Berlin, 10587 Berlin (Germany); Gege, P.; Schwarzmaier, T.; Lenhard, K.; Baumgartner, A. [Deutsches Zentrum fuer Luft- und Raumfahrt, Institut fuer Methodik der Fernerkundung, 82234 Oberpfaffenhofen (Germany)

    2013-05-10

    A dedicated calibration technique was applied for the calibration of the spectral radiance transfer standard (RASTA) of the German Aerospace Center (DLR) at the Physikalisch-Technische Bundesanstalt (PTB), consisting of two independent but complementing calibration procedures to provide redundancy and smallest possible calibration uncertainties. Procedure I included two calibration steps: In a first step the optical radiation source of RASTA, an FEL lamp, was calibrated in terms of its spectral irradiance E{sub {lambda}}({lambda}) in the wavelength range from 350 nm to 2400 nm using the PTB Spectral Irradiance Calibration Equipment (SPICE), while in a second step the spectral radiance factor {beta}{sub 0 Degree-Sign :45 Degree-Sign }({lambda}) of the RASTA reflection standard was calibrated in a 0 Degree-Sign :45 Degree-Sign -viewing geometry in the wavelength range from 350 nm to 1700 nm at the robot-based gonioreflectometer facility of PTB. The achieved relative standard uncertainties (k= 1) range from 0.6 % to 3.2 % and 0.1 % to 0.6 % respectively. Procedure II was completely independent from procedure I and allowed to cover the entire spectral range of RASTA from 350 nm to 2500 nm. In the second procedure, the 0 Degree-Sign :45 Degree-Sign -viewing geometry spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign }({lambda}) of RASTA was directly calibrated at the Spectral Radiance Comparator Facility (SRCF) of PTB. The relative uncertainties for this calibration procedure range from 0.8 % in the visible up to 7.5 % at 2500 nm (k= 1). In the overlapping spectral range of both calibration procedures the calculated spectral radiance L{sub {lambda},0 Degree-Sign :45 Degree-Sign ,calc}({lambda}) from procedure I is in good agreement with the direct measurement of procedure II, i.e. well within the combined expanded uncertainties (k= 2) of both procedures.

  9. Using inductively coupled plasma-mass spectrometry for calibration transfer between environmental CRMs.

    Science.gov (United States)

    Turk, G C; Yu, L L; Salit, M L; Guthrie, W F

    2001-06-01

    Multielement analyses of environmental reference materials have been performed using existing certified reference materials (CRMs) as calibration standards for inductively coupled plasma-mass spectrometry. The analyses have been performed using a high-performance methodology that results in comparison measurement uncertainties that are significantly less than the uncertainties of the certified values of the calibration CRM. Consequently, the determined values have uncertainties that are very nearly equivalent to the uncertainties of the calibration CRM. Several uses of this calibration transfer are proposed, including, re-certification measurements of replacement CRMs, establishing traceability of one CRM to another, and demonstrating the equivalence of two CRMs. RM 8704, a river sediment, was analyzed using SRM 2704, Buffalo River Sediment, as the calibration standard. SRM 1632c, Trace Elements in Bituminous Coal, which is a replacement for SRM 1632b, was analyzed using SRM 1632b as the standard. SRM 1635, Trace Elements in Subbituminous Coal, was also analyzed using SRM 1632b as the standard.

  10. Establishment of a procedure to calculate the measurement uncertainties in radiation survey meters calibration

    International Nuclear Information System (INIS)

    Manzoli, J.E.; Potiens, M.P.A.

    2000-01-01

    The Calibration Laboratory of Sao Paulo calibrates more than one thousand gamma ray survey meters a year; beside other kinds of radiotherapy, radiodiagnostic and radiation protection instruments. It has a standard (600 cm 3 ) cylinder ionization chamber (Nuclear Enterprises Ltd. model 2511/3) traceable to the Brazilian Secondary Standard Dosimetry Laboratory (SSDL) whose instruments are traceable to the BIPM. Annually the beam dosimetry is performed using this chamber and the results are used as the true values for calibration purposes. The uncertainties present in every direct or indirect measurement during the calibration procedure must be evaluated for purposes of laboratory quality control. All calculation steps in the propagation of errors are presented in this work staging from the ionization chamber charge measured with the standard instrument. Such a propagation was made in space and time, considering even the environmental quantities uncertainties. The propagation was necessary in space, because the ionization chamber measurements were performed at only one space position. The time propagation was essential due to the fact that the activity is a peculiar physical quantity which changes with time according to precise relations for a specific radionuclide. The clear indication of every measurement uncertainty is always important to quantify the quality of this measurement. Nowadays the achievement of calibration laboratory quality systems requires the expression of all uncertainties and the procedure used to evaluate it. An example of this procedure in the case of the calibration of a typical portable radiation survey meter is presented. The direct exposure rate instrument measurement was compared with the true value given by the standard instrument properly propagated and all quantities used have their uncertainties shown. (author)

  11. Technical Note: Procedure for the calibration and validation of kilo-voltage cone-beam CT models

    Energy Technology Data Exchange (ETDEWEB)

    Vilches-Freixas, Gloria; Létang, Jean Michel; Rit, Simon, E-mail: simon.rit@creatis.insa-lyon.fr [Université de Lyon, CREATIS, CNRS UMR5220, Inserm U1206, INSA-Lyon, Université Lyon 1, Centre Léon Bérard, Lyon 69373 Cedex 08 (France); Brousmiche, Sébastien [Ion Beam Application, Louvain-la-Neuve 1348 (Belgium); Romero, Edward; Vila Oliva, Marc [Université de Lyon, CREATIS, CNRS UMR5220, Inserm U1206, INSA-Lyon, Université Lyon 1, Centre Léon Bérard, Lyon 69373 Cedex 08, France and Ion Beam Application, Louvain-la-Neuve 1348 (Belgium); Kellner, Daniel; Deutschmann, Heinz; Keuschnigg, Peter; Steininger, Philipp [Institute for Research and Development on Advanced Radiation Technologies, Paracelsus Medical University, Salzburg 5020 (Austria)

    2016-09-15

    Purpose: The aim of this work is to propose a general and simple procedure for the calibration and validation of kilo-voltage cone-beam CT (kV CBCT) models against experimental data. Methods: The calibration and validation of the CT model is a two-step procedure: the source model then the detector model. The source is described by the direction dependent photon energy spectrum at each voltage while the detector is described by the pixel intensity value as a function of the direction and the energy of incident photons. The measurements for the source consist of a series of dose measurements in air performed at each voltage with varying filter thicknesses and materials in front of the x-ray tube. The measurements for the detector are acquisitions of projection images using the same filters and several tube voltages. The proposed procedure has been applied to calibrate and assess the accuracy of simple models of the source and the detector of three commercial kV CBCT units. If the CBCT system models had been calibrated differently, the current procedure would have been exclusively used to validate the models. Several high-purity attenuation filters of aluminum, copper, and silver combined with a dosimeter which is sensitive to the range of voltages of interest were used. A sensitivity analysis of the model has also been conducted for each parameter of the source and the detector models. Results: Average deviations between experimental and theoretical dose values are below 1.5% after calibration for the three x-ray sources. The predicted energy deposited in the detector agrees with experimental data within 4% for all imaging systems. Conclusions: The authors developed and applied an experimental procedure to calibrate and validate any model of the source and the detector of a CBCT unit. The present protocol has been successfully applied to three x-ray imaging systems. The minimum requirements in terms of material and equipment would make its implementation suitable in

  12. Calibration issues of the TFTR multichannel neutron collimator

    International Nuclear Information System (INIS)

    Goeler, S. von; Johnson, L.C.; Bitter, M.; Efthimion, P.C.; Roquemore, A.L.

    1996-01-01

    The calibration procedures for the detectors in the Neutron Collimator are reviewed. The absolute calibration was performed for the NE451 detectors, in situ, by moving a DT neutron generator in the TFTR vacuum vessel across each sight line. This calibration was transferred to other detectors in the same channel. Four new sight lines have been installed at a different toroidal location, which view the plasma through the vacuum vessel port cover rather than through thinned windows. The new detectors are cross-calibrated to the NE451 detectors with a jog shot procedure, where the plasma is quickly shifted in major radius over a distance of 30 cm. The jog shot procedure shows that scattered neutrons account approximately for 30% of the signal of the new central channels. The neutron source strength from the collimator agrees within 10% with the source strength from global neutron monitors in the TFTR test cell. Detector non-linearity is discussed. Another special issue is the behavior of the detectors during T-puffs, where the DD/DT neutron ratio changes rapidly

  13. Calibration of Galileo signals for time metrology.

    Science.gov (United States)

    Defraigne, Pascale; Aerts, Wim; Cerretto, Giancarlo; Cantoni, Elena; Sleewaegen, Jean-Marie

    2014-12-01

    Using global navigation satellite system (GNSS) signals for accurate timing and time transfer requires the knowledge of all electric delays of the signals inside the receiving system. GNSS stations dedicated to timing or time transfer are classically calibrated only for Global Positioning System (GPS) signals. This paper proposes a procedure to determine the hardware delays of a GNSS receiving station for Galileo signals, once the delays of the GPS signals are known. This approach makes use of the broadcast satellite inter-signal biases, and is based on the ionospheric delay measured from dual-frequency combinations of GPS and Galileo signals. The uncertainty on the so-determined hardware delays is estimated to 3.7 ns for each isolated code in the L5 frequency band, and 4.2 ns for the ionosphere-free combination of E1 with a code of the L5 frequency band. For the calibration of a time transfer link between two stations, another approach can be used, based on the difference between the common-view time transfer results obtained with calibrated GPS data and with uncalibrated Galileo data. It is shown that the results obtained with this approach or with the ionospheric method are equivalent.

  14. Calibration Transfer Between a Bench Scanning and a Submersible Diode Array Spectrophotometer for In Situ Wastewater Quality Monitoring in Sewer Systems.

    Science.gov (United States)

    Brito, Rita S; Pinheiro, Helena M; Ferreira, Filipa; Matos, José S; Pinheiro, Alexandre; Lourenço, Nídia D

    2016-03-01

    Online monitoring programs based on spectroscopy have a high application potential for the detection of hazardous wastewater discharges in sewer systems. Wastewater hydraulics poses a challenge for in situ spectroscopy, especially when the system includes storm water connections leading to rapid changes in water depth, velocity, and in the water quality matrix. Thus, there is a need to optimize and fix the location of in situ instruments, limiting their availability for calibration. In this context, the development of calibration models on bench spectrophotometers to estimate wastewater quality parameters from spectra acquired with in situ instruments could be very useful. However, spectra contain information not only from the samples, but also from the spectrophotometer generally invalidating this approach. The use of calibration transfer methods is a promising solution to this problem. In this study, calibration models were developed using interval partial least squares (iPLS), for the estimation of total suspended solids (TSS) and chemical oxygen demand (COD) in sewage from Ultraviolet-visible spectra acquired in a bench scanning spectrophotometer. The feasibility of calibration transfer to a submersible, diode array equipment, to be subsequently operated in situ, was assessed using three procedures: slope and bias correction (SBC); single wavelength standardization (SWS) on mean spectra; and local centering (LC). The results showed that SBC was the most adequate for the available data, adding insignificant error to the base model estimates. Single wavelength standardization was a close second best, potentially more robust, and independent of the base iPLS model. Local centering was shown to be inadequate for the samples and instruments used. © The Author(s) 2016.

  15. Linear model correction: A method for transferring a near-infrared multivariate calibration model without standard samples

    Science.gov (United States)

    Liu, Yan; Cai, Wensheng; Shao, Xueguang

    2016-12-01

    Calibration transfer is essential for practical applications of near infrared (NIR) spectroscopy because the measurements of the spectra may be performed on different instruments and the difference between the instruments must be corrected. For most of calibration transfer methods, standard samples are necessary to construct the transfer model using the spectra of the samples measured on two instruments, named as master and slave instrument, respectively. In this work, a method named as linear model correction (LMC) is proposed for calibration transfer without standard samples. The method is based on the fact that, for the samples with similar physical and chemical properties, the spectra measured on different instruments are linearly correlated. The fact makes the coefficients of the linear models constructed by the spectra measured on different instruments are similar in profile. Therefore, by using the constrained optimization method, the coefficients of the master model can be transferred into that of the slave model with a few spectra measured on slave instrument. Two NIR datasets of corn and plant leaf samples measured with different instruments are used to test the performance of the method. The results show that, for both the datasets, the spectra can be correctly predicted using the transferred partial least squares (PLS) models. Because standard samples are not necessary in the method, it may be more useful in practical uses.

  16. Calibration procedures of the Tore-Supra infrared endoscopes

    Science.gov (United States)

    Desgranges, C.; Jouve, M.; Balorin, C.; Reichle, R.; Firdaouss, M.; Lipa, M.; Chantant, M.; Gardarein, J. L.; Saille, A.; Loarer, T.

    2018-01-01

    Five endoscopes equipped with infrared cameras working in the medium infrared range (3-5 μm) are installed on the controlled thermonuclear fusion research device Tore-Supra. These endoscopes aim at monitoring the plasma facing components surface temperature to prevent their overheating. Signals delivered by infrared cameras through endoscopes are analysed and used on the one hand through a real time feedback control loop acting on the heating systems of the plasma to decrease plasma facing components surface temperatures when necessary, on the other hand for physics studies such as determination of the incoming heat flux . To ensure these two roles a very accurate knowledge of the absolute surface temperatures is mandatory. Consequently the infrared endoscopes must be calibrated through a very careful procedure. This means determining their transmission coefficients which is a delicate operation. Methods to calibrate infrared endoscopes during the shutdown period of the Tore-Supra machine will be presented. As they do not allow determining the possible transmittances evolution during operation an in-situ method is presented. It permits the validation of the calibration performed in laboratory as well as the monitoring of their evolution during machine operation. This is possible by the use of the endoscope shutter and a dedicated plasma scenario developed to heat it. Possible improvements of this method are briefly evoked.

  17. Calibration of the modulation transfer function of surface profilometers with binary pseudo-random test standards: expanding the application range

    International Nuclear Information System (INIS)

    Yashchuk, Valeriy V.; Anderson, Erik H.; Barber, Samuel K.; Bouet, Nathalie; Cambie, Rossana; Conley, Raymond; McKinney, Wayne R.; Takacs, Peter Z.; Voronov, Dmitriy L.

    2011-01-01

    A modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) gratings and arrays (Proc. SPIE 7077-7 (2007), Opt. Eng. 47, 073602 (2008)) has been proven to be an effective MTF calibration method for a number of interferometric microscopes and a scatterometer (Nucl. Instr. and Meth. A616, 172 (2010)). Here we report on a further expansion of the application range of the method. We describe the MTF calibration of a 6 inch phase shifting Fizeau interferometer. Beyond providing a direct measurement of the interferometer's MTF, tests with a BPR array surface have revealed an asymmetry in the instrument's data processing algorithm that fundamentally limits its bandwidth. Moreover, the tests have illustrated the effects of the instrument's detrending and filtering procedures on power spectral density measurements. The details of the development of a BPR test sample suitable for calibration of scanning and transmission electron microscopes are also presented. Such a test sample is realized as a multilayer structure with the layer thicknesses of two materials corresponding to BPR sequence. The investigations confirm the universal character of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  18. Rapid procedure to calibrate EC-10 and EC-20 capacitance sensors ...

    African Journals Online (AJOL)

    A rapid calibration procedure for EC-10 and EC-20 sensors is introduced to promote the commercial use of these sensors for hydroponic irrigation management in coir. The method is comprised of taking one sensor reading, by a sensor installed under hydroponic crop production conditions, and one gravimetric sample, ...

  19. Primary calibration of TXRF in comparison with the standard droplet procedure

    International Nuclear Information System (INIS)

    Dobler, M.; Reus, U.; Knoth, J.; Schwenke, H.

    2000-01-01

    For the determination of contamination on wafer surfaces with total reflection x-ray fluorescence spectrometry (TXRF) normally external 1 ng Ni droplet standards were used for calibration. This method is based on several assumptions about the properties of the standard droplet which are strongly affected by the preparation of the samples. In this paper a study is resented which compares the external droplet method with a calibration procedure using the fundamental physical background of total reflection x-ray fluorescence spectrometry and the properties of Ni bulk material. The particular features of the two methods will be discussed and the obtained results compared to each other. It is demonstrated in this study that the calibration with Ni bulk material is a primary method which offers several advantages compared to the calibration based on droplet standards. These advantages are unique and enable a more reliable and reproducible quantification of contamination on wafer surfaces. This is caused by the fact that the method is standardless and only based on fundamental parameters and natural constants. It is also demonstrated that effects which could be caused by especial features of the measured samples (particle or film, e.g.) or by the degradation of the calibration sample could be excluded. (author)

  20. Offline calibration procedure of the CMS Drift Tube detectors

    International Nuclear Information System (INIS)

    Abbiendi, G; Battilana, C; Cavallo, F R; Giunta, M; Guiducci, L; Amapane, N; Bolognesi, S; Cerminara, G; Bellan, R; Biallass, P; Frangenheim, J; Biasotto, M; Tazon, A Calderon; Cepeda, M; Cruz, B De La; Pardos, C Diez; Bedoya, C Fernandez; Iglesias, M C Fouz; Menendez, J Fernandez; Gresele, A

    2009-01-01

    The barrel region of the Compact Muon Solenoid (CMS) experiment at the Large Hadron Collider is instrumented with Drift Tube (DT) detectors. This paper describes in full details the calibration of the DT hit reconstruction algorithm. After inter-channel synchronization has been verified through the appropriate hardware procedure, the time pedestals are extracted directly from the distribution of the recorded times. Further corrections for time-of-flight and time of signal propagation are applied as soon as the three-dimensional hit position within the DT chamber is known. The different effects of the time pedestal miscalibration on the two main hit reconstruction algorithms are shown. The drift velocity calibration algorithm is based on the meantimer technique. Different meantimer relations for different track angles and patterns of hit cells are used. This algorithm can also be used to determine the uncertainty on the reconstructed hit position.

  1. IAEA workshop/seminar on calibration procedures in dosimetry, Quito, 6-24 October 1986

    International Nuclear Information System (INIS)

    1987-01-01

    The International Atomic Energy Agency in co-operation with the Ecuadorian Atomic Energy Commission organized a workshop and seminar on calibration procedures in dosimetry at the SSDL Quito, 6 to 24 October 1986. All calibration laboratories in the Latin American region were invited to participate. The purpose of the workshop were calibration exercises with therapy-level and protection-level secondary standards at various calibration qualities, discussions on progress made in the different SSDLs in the region and delivering lectures on pertinent subjects. A total of 15 Secondary Standards (10 therapy-level and 5 protection-level) were brought along by the participants and 35 calibration comparisons were performed with those having a valid calibration factor. Thirty-three determinations of calibration factors were performed for secondary standards not having had a calibration before. Twelve different calibration qualities were available (Cobalt-60 and X-rays) and Agency's Secondary Standards traceable to BIPM were the reference standards. The participants were divided into two working groups, each one week and each group into two sub-groups. Both irradiation bunkers were used simultaneously. The one houses the teletherapy Cobalt-60 unit and the protection-level Cobalt-60 irradiator, the other one the constant potential X-ray machine with maximum generating potential of 320 KV and suitable for both therapy-level as well as protection-level calibrations. Due to the heavy workload and limited time available some nightshifts were required to accomplish the requested calibration comparisons

  2. Calibration in atomic spectrometry: A tutorial review dealing with quality criteria, weighting procedures and possible curvatures

    International Nuclear Information System (INIS)

    Mermet, Jean-Michel

    2010-01-01

    Calibration is required to obtain analyte concentrations in atomic spectrometry. To take full benefit of it, the adequacy of the coefficient of determination r 2 is discussed, and its use is compared with the uncertainty due to the prediction bands of the regression. Also discussed from a tutorial point of view are the influence of the weighting procedure and of different weighting factors, and the comparison between linear and quadratic regression to cope with curvatures. They are illustrated with examples based on the use of ICP-AES with nebulization and laser ablation, and of LIBS. Use of a calibration graph over several orders of magnitude may be problematic as well as the use of a quadratic regression to cope with possible curvatures. Instrument softwares that allow reprocessing of the calibration by selecting standards around the expected analyte concentration are convenient for optimizing the calibration procedure.

  3. Discharge and Nitrogen Transfer Modelling in the Berze River: A HYPE Setup and Calibration

    Directory of Open Access Journals (Sweden)

    Veinbergs Arturs

    2017-05-01

    Full Text Available This study is focused on water quality and quantity modelling in the Berze River basin located in the Zemgale region of Latvia. The contributing basin area of 872 km2 is furthermore divided into 15 sub-basins designated according to the characteristics of hydrological network and water sampling programme. The river basin of interest is a spatially complex system with agricultural land and forests as two predominant land use types. Complexity of the system reflects in the discharge intensity and diffuse pollution of nitrogen compounds into the water bodies of the river basin. The presence of urban area has an impact as the load from the existing wastewater treatment plants consist up to 76 % of the total nitrogen load in the Berze River basin. Representative data sets of land cover, agricultural field data base for crop distribution analysis, estimation of crop management, soil type map, digital elevation model, drainage conditions, network of water bodies and point sources were used for the modelling procedures. The semi-distributed hydro chemical model HYPE has a setup to simulate discharge and nitrogen transfer. In order to make the model more robust and appropriate for the current study the data sets previously stated were classified by unifying similar spatially located polygons. The data layers were overlaid and 53 hydrological response units (SLCs were created. Agricultural land consists of 48 SLCs with the details of soils, drainage conditions, crop types, and land management practices. Manual calibration procedure was applied to improve the performance of discharge simulation. Simulated discharge values showed good agreement with the observed values with the Nash-Sutcliffe efficiency of 0.82 and bias of −6.6 %. Manual calibration of parameters related to nitrogen leakage simulation was applied to test the most sensitive parameters.

  4. Discharge and Nitrogen Transfer Modelling in the Berze River: A HYPE Setup and Calibration

    Science.gov (United States)

    Veinbergs, Arturs; Lagzdins, Ainis; Jansons, Viesturs; Abramenko, Kaspars; Sudars, Ritvars

    2017-05-01

    This study is focused on water quality and quantity modelling in the Berze River basin located in the Zemgale region of Latvia. The contributing basin area of 872 km2 is furthermore divided into 15 sub-basins designated according to the characteristics of hydrological network and water sampling programme. The river basin of interest is a spatially complex system with agricultural land and forests as two predominant land use types. Complexity of the system reflects in the discharge intensity and diffuse pollution of nitrogen compounds into the water bodies of the river basin. The presence of urban area has an impact as the load from the existing wastewater treatment plants consist up to 76 % of the total nitrogen load in the Berze River basin. Representative data sets of land cover, agricultural field data base for crop distribution analysis, estimation of crop management, soil type map, digital elevation model, drainage conditions, network of water bodies and point sources were used for the modelling procedures. The semi-distributed hydro chemical model HYPE has a setup to simulate discharge and nitrogen transfer. In order to make the model more robust and appropriate for the current study the data sets previously stated were classified by unifying similar spatially located polygons. The data layers were overlaid and 53 hydrological response units (SLCs) were created. Agricultural land consists of 48 SLCs with the details of soils, drainage conditions, crop types, and land management practices. Manual calibration procedure was applied to improve the performance of discharge simulation. Simulated discharge values showed good agreement with the observed values with the Nash-Sutcliffe efficiency of 0.82 and bias of -6.6 %. Manual calibration of parameters related to nitrogen leakage simulation was applied to test the most sensitive parameters.

  5. An alternative method for calibration of narrow band radiometer using a radiative transfer model

    Energy Technology Data Exchange (ETDEWEB)

    Salvador, J; Wolfram, E; D' Elia, R [Centro de Investigaciones en Laseres y Aplicaciones, CEILAP (CITEFA-CONICET), Juan B. de La Salle 4397 (B1603ALO), Villa Martelli, Buenos Aires (Argentina); Zamorano, F; Casiccia, C [Laboratorio de Ozono y Radiacion UV, Universidad de Magallanes, Punta Arenas (Chile) (Chile); Rosales, A [Universidad Nacional de la Patagonia San Juan Bosco, UNPSJB, Facultad de Ingenieria, Trelew (Argentina) (Argentina); Quel, E, E-mail: jsalvador@citefa.gov.ar [Universidad Nacional de la Patagonia Austral, Unidad Academica Rio Gallegos Avda. Lisandro de la Torre 1070 ciudad de Rio Gallegos-Sta Cruz (Argentina) (Argentina)

    2011-01-01

    The continual monitoring of solar UV radiation is one of the major objectives proposed by many atmosphere research groups. The purpose of this task is to determine the status and degree of progress over time of the anthropogenic composition perturbation of the atmosphere. Such changes affect the intensity of the UV solar radiation transmitted through the atmosphere that then interacts with living organisms and all materials, causing serious consequences in terms of human health and durability of materials that interact with this radiation. One of the many challenges that need to be faced to perform these measurements correctly is the maintenance of periodic calibrations of these instruments. Otherwise, damage caused by the UV radiation received will render any one calibration useless after the passage of some time. This requirement makes the usage of these instruments unattractive, and the lack of frequent calibration may lead to the loss of large amounts of acquired data. Motivated by this need to maintain calibration or, at least, know the degree of stability of instrumental behavior, we have developed a calibration methodology that uses the potential of radiative transfer models to model solar radiation with 5% accuracy or better relative to actual conditions. Voltage values in each radiometer channel involved in the calibration process are carefully selected from clear sky data. Thus, tables are constructed with voltage values corresponding to various atmospheric conditions for a given solar zenith angle. Then we model with a radiative transfer model using the same conditions as for the measurements to assemble sets of values for each zenith angle. The ratio of each group (measured and modeled) allows us to calculate the calibration coefficient value as a function of zenith angle as well as the cosine response presented by the radiometer. The calibration results obtained by this method were compared with those obtained with a Brewer MKIII SN 80 located in the

  6. Calibration procedures for improved accuracy of wind turbine blade load measurement

    Energy Technology Data Exchange (ETDEWEB)

    Dahlberg, J.Aa. [Aeronautical Research Inst. of Sweden, Bromma (Sweden); Johansson, Hjalmar [Teknikgruppen AB, Sollentuna (Sweden)

    1996-12-01

    External loads acting on wind turbine blades are mainly transferred via the hub to the rest of the structure. It is therefore a normal approach to measure the loads acting on the turbine by load measurements in the blade roots. The load measurement is often accomplished by measurements of strain on the surface of the blade or the hub. The strain signals are converted to loads by applying calibration factors to the measurements. This paper deals with difficulties associated with load measurements on two different wind turbines; one with strain gauges applied to a steel hub where a linear stress-load relationship is expected and the other with strain gauges applied to the GFRP blade close to the bearings where strong non-linearity`s and temperature effects are expected. This paper suggests calibration methods to overcome these problems. 2 refs, 11 figs

  7. A procedure to validate and correct the {sup 13}C chemical shift calibration of RNA datasets

    Energy Technology Data Exchange (ETDEWEB)

    Aeschbacher, Thomas; Schubert, Mario, E-mail: schubert@mol.biol.ethz.ch; Allain, Frederic H.-T., E-mail: allain@mol.biol.ethz.ch [ETH Zuerich, Institute for Molecular Biology and Biophysics (Switzerland)

    2012-02-15

    Chemical shifts reflect the structural environment of a certain nucleus and can be used to extract structural and dynamic information. Proper calibration is indispensable to extract such information from chemical shifts. Whereas a variety of procedures exist to verify the chemical shift calibration for proteins, no such procedure is available for RNAs to date. We present here a procedure to analyze and correct the calibration of {sup 13}C NMR data of RNAs. Our procedure uses five {sup 13}C chemical shifts as a reference, each of them found in a narrow shift range in most datasets deposited in the Biological Magnetic Resonance Bank. In 49 datasets we could evaluate the {sup 13}C calibration and detect errors or inconsistencies in RNA {sup 13}C chemical shifts based on these chemical shift reference values. More than half of the datasets (27 out of those 49) were found to be improperly referenced or contained inconsistencies. This large inconsistency rate possibly explains that no clear structure-{sup 13}C chemical shift relationship has emerged for RNA so far. We were able to recalibrate or correct 17 datasets resulting in 39 usable {sup 13}C datasets. 6 new datasets from our lab were used to verify our method increasing the database to 45 usable datasets. We can now search for structure-chemical shift relationships with this improved list of {sup 13}C chemical shift data. This is demonstrated by a clear relationship between ribose {sup 13}C shifts and the sugar pucker, which can be used to predict a C2 Prime - or C3 Prime -endo conformation of the ribose with high accuracy. The improved quality of the chemical shift data allows statistical analysis with the potential to facilitate assignment procedures, and the extraction of restraints for structure calculations of RNA.

  8. pH sensor calibration procedure

    OpenAIRE

    Artero Delgado, Carola; Nogueras Cervera, Marc; Manuel Lázaro, Antonio; Prat Tasias, Jordi; Prat Farran, Joana d'Arc

    2013-01-01

    This paper describes the calibration of pH sensor located at the OBSEA marine Observatory. This instrument is based on an industrial pH electrode that is connected to a CTD instrument (Conductivity, Temperature, and Depth ). The calibration of the pH sensor has been done using a high precision spectrophotometer pH meter from Institute of Marine Sciences (ICM), and in this way it has been obtained a numerical function for the p H sensor propor...

  9. Procedure for the Selection and Validation of a Calibration Model I-Description and Application.

    Science.gov (United States)

    Desharnais, Brigitte; Camirand-Lemyre, Félix; Mireault, Pascal; Skinner, Cameron D

    2017-05-01

    Calibration model selection is required for all quantitative methods in toxicology and more broadly in bioanalysis. This typically involves selecting the equation order (quadratic or linear) and weighting factor correctly modelizing the data. A mis-selection of the calibration model will generate lower quality control (QC) accuracy, with an error up to 154%. Unfortunately, simple tools to perform this selection and tests to validate the resulting model are lacking. We present a stepwise, analyst-independent scheme for selection and validation of calibration models. The success rate of this scheme is on average 40% higher than a traditional "fit and check the QCs accuracy" method of selecting the calibration model. Moreover, the process was completely automated through a script (available in Supplemental Data 3) running in RStudio (free, open-source software). The need for weighting was assessed through an F-test using the variances of the upper limit of quantification and lower limit of quantification replicate measurements. When weighting was required, the choice between 1/x and 1/x2 was determined by calculating which option generated the smallest spread of weighted normalized variances. Finally, model order was selected through a partial F-test. The chosen calibration model was validated through Cramer-von Mises or Kolmogorov-Smirnov normality testing of the standardized residuals. Performance of the different tests was assessed using 50 simulated data sets per possible calibration model (e.g., linear-no weight, quadratic-no weight, linear-1/x, etc.). This first of two papers describes the tests, procedures and outcomes of the developed procedure using real LC-MS-MS results for the quantification of cocaine and naltrexone. © The Author 2017. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

  10. Calibration of Modulation Transfer Function of Surface Profilometers with 1D and 2D Binary Pseudo-random Array Standards

    International Nuclear Information System (INIS)

    Yashchuk, Valeriy V.; McKinney, Wayne R.; Takacs, Peter Z.

    2008-01-01

    We suggest and describe the use of a binary pseudo-random grating as a standard test surface for calibration of the modulation transfer function of microscopes. Results from calibration of a MicromapTM-570 interferometric microscope are presented.

  11. Enhanced Single Seed Trait Predictions in Soybean (Glycine max) and Robust Calibration Model Transfer with Near-Infrared Reflectance Spectroscopy.

    Science.gov (United States)

    Hacisalihoglu, Gokhan; Gustin, Jeffery L; Louisma, Jean; Armstrong, Paul; Peter, Gary F; Walker, Alejandro R; Settles, A Mark

    2016-02-10

    Single seed near-infrared reflectance (NIR) spectroscopy predicts soybean (Glycine max) seed quality traits of moisture, oil, and protein. We tested the accuracy of transferring calibrations between different single seed NIR analyzers of the same design by collecting NIR spectra and analytical trait data for globally diverse soybean germplasm. X-ray microcomputed tomography (μCT) was used to collect seed density and shape traits to enhance the number of soybean traits that can be predicted from single seed NIR. Partial least-squares (PLS) regression gave accurate predictive models for oil, weight, volume, protein, and maximal cross-sectional area of the seed. PLS models for width, length, and density were not predictive. Although principal component analysis (PCA) of the NIR spectra showed that black seed coat color had significant signal, excluding black seeds from the calibrations did not impact model accuracies. Calibrations for oil and protein developed in this study as well as earlier calibrations for a separate NIR analyzer of the same design were used to test the ability to transfer PLS regressions between platforms. PLS models built from data collected on one NIR analyzer had minimal differences in accuracy when applied to spectra collected from a sister device. Model transfer was more robust when spectra were trimmed from 910 to 1679 nm to 955-1635 nm due to divergence of edge wavelengths between the two devices. The ability to transfer calibrations between similar single seed NIR spectrometers facilitates broader adoption of this high-throughput, nondestructive, seed phenotyping technology.

  12. Radon in air calibration procedure: A primary method

    International Nuclear Information System (INIS)

    Lucas, H.F.; Markun, F.

    1988-01-01

    A procedure has been developed for preparing 3- to 9-iota volumes of air under natural conditions with a known concentration of /sup 222/Rn to be used for calibrating radon systems. Air is passed into a plastic bag through a standard /sup 226/Ra solution (prepared by the U.S. National Bureau of Standards) contained in an emanation flask. This plastic bag retains /sup 222/Rn with little loss into or through the bag walls. The mean ratios of the /sup 222/Rn in the air at 2 and 7 days after filling to that immediately after filling were -.992 +- 0.006 and 0.969 +- 0.008, which suggests a rate of radon loss into the bag of 0.4 +- 0.1%/day

  13. A versatile calibration procedure for portable coded aperture gamma cameras and RGB-D sensors

    Science.gov (United States)

    Paradiso, V.; Crivellaro, A.; Amgarou, K.; de Lanaute, N. Blanc; Fua, P.; Liénard, E.

    2018-04-01

    The present paper proposes a versatile procedure for the geometrical calibration of coded aperture gamma cameras and RGB-D depth sensors, using only one radioactive point source and a simple experimental set-up. Calibration data is then used for accurately aligning radiation images retrieved by means of the γ-camera with the respective depth images computed with the RGB-D sensor. The system resulting from such a combination is thus able to retrieve, automatically, the distance of radioactive hotspots by means of pixel-wise mapping between gamma and depth images. This procedure is of great interest for a wide number of applications, ranging from precise automatic estimation of the shape and distance of radioactive objects to Augmented Reality systems. Incidentally, the corresponding results validated the choice of a perspective design model for a coded aperture γ-camera.

  14. A general procedure for thermomechanical calibration of nano/micro-mechanical resonators

    International Nuclear Information System (INIS)

    Hauer, B.D.; Doolin, C.; Beach, K.S.D.; Davis, J.P.

    2013-01-01

    We describe a general procedure to calibrate the detection of a nano/micro-mechanical resonator’s displacement as it undergoes thermal Brownian motion. A brief introduction to the equations of motion for such a resonator is presented, followed by a detailed derivation of the corresponding power spectral density (PSD) function, which is identical in all situations aside from a system-dependent effective mass value. The effective masses for a number of different resonator geometries are determined using both finite element method (FEM) modeling and analytical calculations. -- Highlights: •Model micro- and nanomechanical resonators displaced by their own thermal motion. •Review the theoretical framework for describing thermomechanical systems. •Present a recipe for measurement calibration on devices of arbitrary shape. •Point out and correct inconsistencies in the existing literature. •Provide an authoritative guide and reference for practitioners in this area

  15. A general procedure for thermomechanical calibration of nano/micro-mechanical resonators

    Energy Technology Data Exchange (ETDEWEB)

    Hauer, B.D., E-mail: bhauer@ualberta.ca; Doolin, C.; Beach, K.S.D., E-mail: kbeach@ualberta.ca; Davis, J.P., E-mail: jdavis@ualberta.ca

    2013-12-15

    We describe a general procedure to calibrate the detection of a nano/micro-mechanical resonator’s displacement as it undergoes thermal Brownian motion. A brief introduction to the equations of motion for such a resonator is presented, followed by a detailed derivation of the corresponding power spectral density (PSD) function, which is identical in all situations aside from a system-dependent effective mass value. The effective masses for a number of different resonator geometries are determined using both finite element method (FEM) modeling and analytical calculations. -- Highlights: •Model micro- and nanomechanical resonators displaced by their own thermal motion. •Review the theoretical framework for describing thermomechanical systems. •Present a recipe for measurement calibration on devices of arbitrary shape. •Point out and correct inconsistencies in the existing literature. •Provide an authoritative guide and reference for practitioners in this area.

  16. Experimental procedures for the calibration of scintillation cells used in the determination of radon gas concentrations

    International Nuclear Information System (INIS)

    Grenier, M; Bigu, J.

    1982-02-01

    Experimental and analytical procedures are described for the calibration of scintillation cells used for the determination of radon gas concentration. In-house designed and built scintillation cells, used routinely in the monitoring of radon gas in uranium mine underground environments and in the laboratory, were calibrated. The cells had a volume of approximately 158 cm 3 and an α-counting efficiency ranging from 50% to 64%. Calibration factors for the cells were determined. Values ranged approximately from 0.177 cpm/pCiL -1 (4.77 cpm/BqL -1 ) to 0.224 cpm/pCiL -1 (6.05 cpm/BqL -1 ). The calibration facilities at the Elliot Lake Laboratory are briefly described

  17. High voltage calibration of the TANSY-KM5 neutron detectors

    International Nuclear Information System (INIS)

    Grosshoeg, G.; Belle, P. van; Wilson, D.

    1996-11-01

    We have developed a procedure for the high voltage calibration of the TANSY neutron detectors. The procedure is based on the work done during the construction of the spectrometer. A program is written for the measurement of the sensitivity of the neutron detectors as a function of the high voltage. The data are transferred to a PC for evaluation. We use a Cobalt source for the calibration. With the PC the voltage corresponding to the effective Compton edge is found. The voltage settings for the neutron detectors are calculated and stored in a file suitable for input to a program that is used to control the instrument. A measurement is reported that shows that the reproducibility of the measurement is good. 4 refs

  18. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities

    International Nuclear Information System (INIS)

    Costa, Alessandro Martins da

    1999-01-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  19. Simplified calibration and evaluation procedures for improvised whole body gamma spectrometry in emergency situations

    International Nuclear Information System (INIS)

    Malatova, I.; Bucina, I.; Drabova, D.; Cespirova, I.

    2000-01-01

    A semiconductor gamma spectrometer could be used for the rapid estimation of internal contamination of the people in cases of accidents even when no special calibration for whole body counting is prepared. Generic transfer factors for calculation of the whole body detection efficiency from the 25 cm distant point source detection efficiency are presented. Generic dependence of parameters of power function describing the detector efficiency for point source in 25 cm on the detector relative efficiency given by producer was derived from calibration of 18 detectors with relative efficiency from 1.4% to 62%. Minimum detectable activity for various backgrounds and the uncertainty of the estimate of whole body retention are presented, too. (author)

  20. Intercomparison of calibration procedures of high dose rate 192 Ir sources in Brazil and a proposal of a new methodology

    International Nuclear Information System (INIS)

    Marechal, M.H.; Almeida, C.E. de

    1998-01-01

    The objective of this paper is to report the results of an intercomparison of the calibration procedures for 192 Ir sources presently in use in Brazil and to proposal a calibration procedure to derive the N k for a Farmer type ionization chamber for 192 Ir energy by interpolating from a 60 Co gamma-rays and 250 kV x-rays calibration factors. the intercomparison results were all within ± 3.0 % except one case where 4.6 % was observed and latter identified as a problem with N-k value for X-rays. The method proposed by the present work make possible the improvement of the metrological coherence among the calibration laboratories and their users once the N k values could then provided by any of the members of SSDL network. (Author)

  1. Experimental comparison between total calibration factors and components calibration factors of reference dosemeters used in secondary standard laboratory dosemeters

    International Nuclear Information System (INIS)

    Silva, T.A. da.

    1981-06-01

    A quantitative comparison of component calibration factors with the corresponding overall calibration factor was used to evaluate the adopted component calibration procedure in regard to parasitic elements. Judgement of significance is based upon the experimental uncertainty of a well established procedure for determination of the overall calibration factor. The experimental results obtained for different ionization chambers and different electrometers demonstrate that for one type of electrometer the parasitic elements have no influence on its sensitivity considering the experimental uncertainty of the calibration procedures. In this case the adopted procedure for determination of component calibration factors is considered to be equivalent to the procedure of determination of the overall calibration factor and thus might be used as a strong quality control measure in routine calibration. (Author) [pt

  2. Procedures for calibration of clinical dosemeters of radiotherapy with gamma radiation Co-60 in the IPEN-CNEN, Sao Paulo, Brazil

    International Nuclear Information System (INIS)

    Damatto, Willian B.; Santos, Gelson P.; Potiens, Maria da Penha A.; Vivolo, Vitor

    2009-01-01

    The periodic calibration of clinic dosimeters is a very important task, because those measurement systems performing the control quality of the radioactive source and for that assure the control quality of the patient dose estimative. This paper shows the calibration procedures and routine improved in the calibration laboratory of IPEN/CNEN-SP to be applied to the clinic dosimeters following international and national recommendations and yet the spend care that the clients should have with the utilization of that measurement systems for control quality of the radiotherapy procedures, for reduce the dose received for patient, and workers, and for increase the confidence level of the measurements.(author)

  3. A recommendation for revised dose calibrator measurement procedures for 89Zr and 124I.

    Directory of Open Access Journals (Sweden)

    Bradley J Beattie

    Full Text Available Because of their chemical properties and multiday half lives, iodine-124 and zirconium-89 are being used in a growing number of PET imaging studies. Some aspects of their quantitation, however, still need attention. For (89Zr the PET images should, in principle, be as quantitatively accurate as similarly reconstructed 18F measurements. We found, however, that images of a 20 cm well calibration phantom containing (89Zr underestimated the activity by approximately 10% relative to a dose calibrator measurement (Capintec CRC-15R using a published calibration setting number of 465. PET images of (124I, in contrast, are complicated by the contribution of decays in cascade that add spurious coincident events to the PET data. When these cascade coincidences are properly accounted for, quantitatively accurate images should be possible. We found, however, that even with this correction we still encountered what appeared to be a large variability in the accuracy of the PET images when compared to dose calibrator measurements made using the calibration setting number, 570, recommended by Capintec. We derive new calibration setting numbers for (89Zr and (124I based on their 511 keV photon peaks as measured on an HPGe detector. The peaks were calibrated relative to an 18F standard, the activity level of which was precisely measured in a dose calibrator under well-defined measurement conditions. When measuring (89Zr on a Capintec CRC-15R we propose the use of calibration setting number 517. And for (124I, we recommend the use of a copper filter surrounding the sample and the use of calibration setting number 494. The new dose calibrator measurement procedures we propose will result in more consistent and accurate radioactivity measurements of (89Zr and (124I. These and other positron emitting radionuclides can be accurately calibrated relative to 18F based on measurements of their 511 keV peaks and knowledge of their relative positron abundances.

  4. Design, calibration and error analysis of instrumentation for heat transfer measurements in internal combustion engines

    Science.gov (United States)

    Ferguson, C. R.; Tree, D. R.; Dewitt, D. P.; Wahiduzzaman, S. A. H.

    1987-01-01

    The paper reports the methodology and uncertainty analyses of instrumentation for heat transfer measurements in internal combustion engines. Results are presented for determining the local wall heat flux in an internal combustion engine (using a surface thermocouple-type heat flux gage) and the apparent flame-temperature and soot volume fraction path length product in a diesel engine (using two-color pyrometry). It is shown that a surface thermocouple heat transfer gage suitably constructed and calibrated will have an accuracy of 5 to 10 percent. It is also shown that, when applying two-color pyrometry to measure the apparent flame temperature and soot volume fraction-path length, it is important to choose at least one of the two wavelengths to lie in the range of 1.3 to 2.3 micrometers. Carefully calibrated two-color pyrometer can ensure that random errors in the apparent flame temperature and in the soot volume fraction path length will remain small (within about 1 percent and 10-percent, respectively).

  5. Calibration

    International Nuclear Information System (INIS)

    Greacen, E.L.; Correll, R.L.; Cunningham, R.B.; Johns, G.G.; Nicolls, K.D.

    1981-01-01

    Procedures common to different methods of calibration of neutron moisture meters are outlined and laboratory and field calibration methods compared. Gross errors which arise from faulty calibration techniques are described. The count rate can be affected by the dry bulk density of the soil, the volumetric content of constitutional hydrogen and other chemical components of the soil and soil solution. Calibration is further complicated by the fact that the neutron meter responds more strongly to the soil properties close to the detector and source. The differences in slope of calibration curves for different soils can be as much as 40%

  6. A computational procedure for finding multiple solutions of convective heat transfer equations

    International Nuclear Information System (INIS)

    Mishra, S; DebRoy, T

    2005-01-01

    In recent years numerical solutions of the convective heat transfer equations have provided significant insight into the complex materials processing operations. However, these computational methods suffer from two major shortcomings. First, these procedures are designed to calculate temperature fields and cooling rates as output and the unidirectional structure of these solutions preclude specification of these variables as input even when their desired values are known. Second, and more important, these procedures cannot determine multiple pathways or multiple sets of input variables to achieve a particular output from the convective heat transfer equations. Here we propose a new method that overcomes the aforementioned shortcomings of the commonly used solutions of the convective heat transfer equations. The procedure combines the conventional numerical solution methods with a real number based genetic algorithm (GA) to achieve bi-directionality, i.e. the ability to calculate the required input variables to achieve a specific output such as temperature field or cooling rate. More important, the ability of the GA to find a population of solutions enables this procedure to search for and find multiple sets of input variables, all of which can lead to the desired specific output. The proposed computational procedure has been applied to convective heat transfer in a liquid layer locally heated on its free surface by an electric arc, where various sets of input variables are computed to achieve a specific fusion zone geometry defined by an equilibrium temperature. Good agreement is achieved between the model predictions and the independent experimental results, indicating significant promise for the application of this procedure in finding multiple solutions of convective heat transfer equations

  7. Multiple-Objective Stepwise Calibration Using Luca

    Science.gov (United States)

    Hay, Lauren E.; Umemoto, Makiko

    2007-01-01

    This report documents Luca (Let us calibrate), a multiple-objective, stepwise, automated procedure for hydrologic model calibration and the associated graphical user interface (GUI). Luca is a wizard-style user-friendly GUI that provides an easy systematic way of building and executing a calibration procedure. The calibration procedure uses the Shuffled Complex Evolution global search algorithm to calibrate any model compiled with the U.S. Geological Survey's Modular Modeling System. This process assures that intermediate and final states of the model are simulated consistently with measured values.

  8. Procedure for calibrating the Technicon Colorimeter I.

    Science.gov (United States)

    Black, J C; Furman, W B

    1975-05-01

    We describe a rapid method for calibrating the Technicon AutoAnalyzer colorimeter I. Test solutions of bromphenol blue are recommended for the calibration, in preference to solutions of potassium dichromate, based on considerations of the instrument's working range and of the stray light characteristics of the associated filters.

  9. SPRT Calibration Uncertainties and Internal Quality Control at a Commercial SPRT Calibration Facility

    Science.gov (United States)

    Wiandt, T. J.

    2008-06-01

    The Hart Scientific Division of the Fluke Corporation operates two accredited standard platinum resistance thermometer (SPRT) calibration facilities, one at the Hart Scientific factory in Utah, USA, and the other at a service facility in Norwich, UK. The US facility is accredited through National Voluntary Laboratory Accreditation Program (NVLAP), and the UK facility is accredited through UKAS. Both provide SPRT calibrations using similar equipment and procedures, and at similar levels of uncertainty. These uncertainties are among the lowest available commercially. To achieve and maintain low uncertainties, it is required that the calibration procedures be thorough and optimized. However, to minimize customer downtime, it is also important that the instruments be calibrated in a timely manner and returned to the customer. Consequently, subjecting the instrument to repeated calibrations or extensive repeated measurements is not a viable approach. Additionally, these laboratories provide SPRT calibration services involving a wide variety of SPRT designs. These designs behave differently, yet predictably, when subjected to calibration measurements. To this end, an evaluation strategy involving both statistical process control and internal consistency measures is utilized to provide confidence in both the instrument calibration and the calibration process. This article describes the calibration facilities, procedure, uncertainty analysis, and internal quality assurance measures employed in the calibration of SPRTs. Data will be reviewed and generalities will be presented. Finally, challenges and considerations for future improvements will be discussed.

  10. Force Measurement Services at Kebs: AN Overview of Equipment, Procedures and Uncertainty

    Science.gov (United States)

    Bangi, J. O.; Maranga, S. M.; Nganga, S. P.; Mutuli, S. M.

    This paper describes the facilities, instrumentation and procedures currently used in the force laboratory at the Kenya Bureau of Standards (KEBS) for force measurement services. The laboratory uses the Force Calibration Machine (FCM) to calibrate force-measuring instruments. The FCM derives its traceability via comparisons using reference transfer force transducers calibrated by the Force Standard Machines (FSM) of a National Metrology Institute (NMI). The force laboratory is accredited to ISO/IEC 17025 by the Germany Accreditation Body (DAkkS). The accredited measurement scope of the laboratory is 1 MN to calibrate force transducers in both compression and tension modes. ISO 376 procedures are used while calibrating force transducers. The KEBS reference transfer standards have capacities of 10, 50, 300 and 1000 kN to cover the full range of the FCM. The uncertainty in the forces measured by the FCM were reviewed and determined in accordance to the new EURAMET calibration guide. The relative expanded uncertainty of force W realized by FCM was evaluated in a range from 10 kN-1 MN, and was found to be 5.0 × 10-4 with the coverage factor k being equal to 2. The overall normalized error (En) of the comparison results was also found to be less than 1. The accredited Calibration and Measurement Capability (CMC) of the KEBS force laboratory was based on the results of those intercomparisons. The FCM enables KEBS to provide traceability for the calibration of class ‘1’ force instruments as per the ISO 376.

  11. Fuel Transfer Cask; Procedure Option and Radiation Protection during Transferring the Spent Fuel

    International Nuclear Information System (INIS)

    Muhammad Khairul Ariff Mustafa; Nurhayati Ramli; Ahmad Nabil Abdul Rahim; Mohd Fazli Zakaria

    2011-01-01

    Reactor TRIGA PUSPATI (RTP) has been operating almost 30 years. Many components are ageing. Nuclear Malaysia has taken an initiative to manage this ageing problem to prolong the life of the reactor. Hence, reactor upgrading project already commence started with the reactor console. To upgrade the core, all the fuel must be taken out from the core. A conceptual design of fuel transfer cask already done. This paper will discuss about the option of safe working procedure for transferring the fuel to the spent fuel pool for temporary. Hence, radiation protection for operator should be considered during the process. (author)

  12. A single model procedure for estimating tank calibration equations

    International Nuclear Information System (INIS)

    Liebetrau, A.M.

    1997-10-01

    A fundamental component of any accountability system for nuclear materials is a tank calibration equation that relates the height of liquid in a tank to its volume. Tank volume calibration equations are typically determined from pairs of height and volume measurements taken in a series of calibration runs. After raw calibration data are standardized to a fixed set of reference conditions, the calibration equation is typically fit by dividing the data into several segments--corresponding to regions in the tank--and independently fitting the data for each segment. The estimates obtained for individual segments must then be combined to obtain an estimate of the entire calibration function. This process is tedious and time-consuming. Moreover, uncertainty estimates may be misleading because it is difficult to properly model run-to-run variability and between-segment correlation. In this paper, the authors describe a model whose parameters can be estimated simultaneously for all segments of the calibration data, thereby eliminating the need for segment-by-segment estimation. The essence of the proposed model is to define a suitable polynomial to fit to each segment and then extend its definition to the domain of the entire calibration function, so that it (the entire calibration function) can be expressed as the sum of these extended polynomials. The model provides defensible estimates of between-run variability and yields a proper treatment of between-segment correlations. A portable software package, called TANCS, has been developed to facilitate the acquisition, standardization, and analysis of tank calibration data. The TANCS package was used for the calculations in an example presented to illustrate the unified modeling approach described in this paper. With TANCS, a trial calibration function can be estimated and evaluated in a matter of minutes

  13. NIST high-dose calibration services

    International Nuclear Information System (INIS)

    Humphreys, J.C.

    1989-01-01

    There is a need for the standardization of high-dose measurements used in the radiation-processing industry in order to provide assured traceability to national standards. NIST provides dosimetry calibration services to this industry. One of these services involves administration of known absorbed doses of gamma rays to customer-supplied dosimeters. The dosimeters are packaged to provide electron equilibrium conditions and are irradiated in a standard 60 Co calibration facility; this provides a calibration of that batch of dosimeters. Another service consists of supplying to a customer calibrated transfer dosimeters for irradiation with the customer's radiation source. The irradiated transfer dosimeters are then returned to NIST for analysis; the results are reported to the customer, providing a calibration of the dose rate of the customer's source. (orig.)

  14. Calibration procedure for a potato crop growth model using information from across Europe

    DEFF Research Database (Denmark)

    Heidmann, Tove; Tofteng, Charlotte; Abrahamsen, Per

    2008-01-01

    for adaptation of the Daisy model to new potato varieties or for the improvement of the existing parameter set. The procedure is then, as a starting point, to focus the calibration process on the recommended list of parameters to change. We demonstrate this approach by showing the procedure for recalibrating...... three varieties using all relevant data from the sites. We believe these new parameterisations to be more robust, because they indirectly were based on information from the six different sites. We claim that this procedure combines both local and specific modeller expertise in a way that results in more......In the FertOrgaNic EU project, 3 years of field experiments with drip irrigation and fertigation were carried out at six different sites across Europe, involving seven different varieties of potato. The Daisy model, which simulates plant growth together with water and nitrogen dynamics, was used...

  15. Transference patron for the calibration of activemeters for use in nuclear medicine

    International Nuclear Information System (INIS)

    Cortes P, A.; Garcia D, O.C.; Ortiz P, I.; Becerril V, A.

    2000-01-01

    With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some μ Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some μ Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm 3 contained in polyethylene small bottles of 25 cm 3 elaborated in the LPR and whose activity is measured with respect to the National Patron ININ-PNM-2. The sources of mid-large life are ampoules with 5 cm 3 of radioactive solution, calibrated in activity by the primary laboratory, LNHB (formerly LMRI) of France. (Author)

  16. Calibration procedures of area monitors in terms of the Ambient Dose Equivalent H*(10), for gamma, x-ray radiation fields

    International Nuclear Information System (INIS)

    Dieguez Davila, L.E.

    1998-01-01

    In the present thesis procedures for calibrating portable survey meters in terms of the new ICRU quantities H*(10) ambient dose equivalent are discussed. Also the remendations of International Comission on Radiation Protection in their report ICRP 60 that inludes the operational magnitudes that the International Comission of Radiation Units proposed for calibrating area monitors

  17. Single-source gamma radiation procedures for improved calibration and measurements in porous media

    International Nuclear Information System (INIS)

    Oostrom, M.; Hofstee, C.; Dane, H.; Lenhard, R.J.

    1998-01-01

    When dual-energy gamma radiation systems are employed for measurements in porous media, count rates from both sources are often used to compute parameter values. However, for several applications, the count rates of just one source are insufficient. These applications include the determination of volumetric liquid content values in two-liquid systems and salt concentration values in water-saturated porous media. Single-energy gamma radiation procedures for three applications are described in this paper. Through an error analysis, single-source procedures are shown to reduce the probable error in the determinations considerably. Example calculations and simple column experiments were conducted for each application to compare the performance of the new single-source and standard dual-source methods. In all cases, the single-source methods provided more reliable data than the traditional dual-source methods. In addition, a single-source calibration procedure is proposed to determine incident count rates indirectly. This procedure, which requires packing under saturated conditions, can be used in all single- and dual-source applications and yields accurate porosity and dry bulk density values

  18. Experiences of a secondary laboratory of dosimetric calibration from the radiation protection and hygiene center CPHR in its first year of work and the procedures for quality assessment used in the calibration and quality control service

    International Nuclear Information System (INIS)

    Morales, J.A.; Campa, R.; Jova Sed, L.

    1996-01-01

    Experiences of a secondary laboratory of dosimetric calibration from the Radiation Protection and Hygiene Center (CPHR) in first year of work and the procedures for quality assessment used in the calibration and quality control service of radiotherapeutic equipment. For the yield calibration of the calibrated sources an ionometric method was used using ionizing chambers coupled to electrometers. Those determination were based on dosimetric American Association of Physicists in Medicine (AAPM)

  19. Rotary mode system initial instrument calibration

    International Nuclear Information System (INIS)

    Johns, B.R.

    1994-01-01

    The attached report contains the vendor calibration procedures used for the initial instrument calibration of the rotary core sampling equipment. The procedures are from approved vendor information files

  20. Automated calibration of TECAN genesis liquid handling workstation utilizing an online balance and density meter.

    Science.gov (United States)

    Xie, Iris H; Wang, Michael H; Carpenter, Richard; Wu, Henry Y

    2004-02-01

    With robotics widely used in bioanalytical assays, accurate system performance is essential to ensure the quality and productivity of the robotics. In our lab, an automated calibration procedure has been developed to evaluate the precision and accuracy of the TECAN (Research Triangle Park, NC, U.S.A.) Genesis liquid handling system in a bioanalytical laboratory setting. The calibrations were performed by transferring and weighing the solvents automatically on a microbalance controlled by a Gemini program. From the data acquired, calibration reports were generated using a template. The novel aspect of this approach is the use of an on-line balance and a density meter, both of which combine to make the calibration process simple, efficient, and precise. For quantitative bioanalysis, a variety of solvents, including methanol, water, mixed solvents, and plasma, are typically used to prepare standards and unknown samples. Density information is usually unknown for the mixed solvents, and the density of plasma can vary from species to species. However, with the use of a universal density meter, the density could be obtained in seconds. The issue of solvent evaporation during the calibration process was also addressed. Calibration curves were set up for various liquid classes. Pipetting volumes ranged from 10 microL to 900 microL. Precision and accuracy results obtained from the semiannual performance evaluations showed this procedure to be reliable and user-friendly. Using the automated calibration procedure, the calibration and performance evaluation of the robotic system is considerably more efficient, and the incidence of unacceptable precision and accuracy is greatly reduced.

  1. Calibration transfer of a Raman spectroscopic quantification method for the assessment of liquid detergent compositions between two at-line instruments installed at two liquid detergent production plants.

    Science.gov (United States)

    Brouckaert, D; Uyttersprot, J-S; Broeckx, W; De Beer, T

    2017-09-01

    Calibration transfer of partial least squares (PLS) quantification models is established between two Raman spectrometers located at two liquid detergent production plants. As full recalibration of existing calibration models is time-consuming, labour-intensive and costly, it is investigated whether the use of mathematical correction methods requiring only a handful of standardization samples can overcome the dissimilarities in spectral response observed between both measurement systems. Univariate and multivariate standardization approaches are investigated, ranging from simple slope/bias correction (SBC), local centring (LC) and single wavelength standardization (SWS) to more complex direct standardization (DS) and piecewise direct standardization (PDS). The results of these five calibration transfer methods are compared reciprocally, as well as with regard to a full recalibration. Four PLS quantification models, each predicting the concentration of one of the four main ingredients in the studied liquid detergent composition, are aimed at transferring. Accuracy profiles are established from the original and transferred quantification models for validation purposes. A reliable representation of the calibration models performance before and after transfer is thus established, based on β-expectation tolerance intervals. For each transferred model, it is investigated whether every future measurement that will be performed in routine will be close enough to the unknown true value of the sample. From this validation, it is concluded that instrument standardization is successful for three out of four investigated calibration models using multivariate (DS and PDS) transfer approaches. The fourth transferred PLS model could not be validated over the investigated concentration range, due to a lack of precision of the slave instrument. Comparing these transfer results to a full recalibration on the slave instrument allows comparison of the predictive power of both Raman

  2. Energy calibration issues in nuclear resonant vibrational spectroscopy: observing small spectral shifts and making fast calibrations.

    Science.gov (United States)

    Wang, Hongxin; Yoda, Yoshitaka; Dong, Weibing; Huang, Songping D

    2013-09-01

    The conventional energy calibration for nuclear resonant vibrational spectroscopy (NRVS) is usually long. Meanwhile, taking NRVS samples out of the cryostat increases the chance of sample damage, which makes it impossible to carry out an energy calibration during one NRVS measurement. In this study, by manipulating the 14.4 keV beam through the main measurement chamber without moving out the NRVS sample, two alternative calibration procedures have been proposed and established: (i) an in situ calibration procedure, which measures the main NRVS sample at stage A and the calibration sample at stage B simultaneously, and calibrates the energies for observing extremely small spectral shifts; for example, the 0.3 meV energy shift between the 100%-(57)Fe-enriched [Fe4S4Cl4](=) and 10%-(57)Fe and 90%-(54)Fe labeled [Fe4S4Cl4](=) has been well resolved; (ii) a quick-switching energy calibration procedure, which reduces each calibration time from 3-4 h to about 30 min. Although the quick-switching calibration is not in situ, it is suitable for normal NRVS measurements.

  3. Calibration of triaxial fluxgate gradiometer

    International Nuclear Information System (INIS)

    Vcelak, Jan

    2006-01-01

    The description of simple and fast calibration procedures used for double-probe triaxial fluxgate gradiometer is provided in this paper. The calibration procedure consists of three basic steps. In the first step both probes are calibrated independently in order to reach constant total field reading in every position. Both probes are numerically aligned in the second step in order that the gradient reading is zero in homogenous magnetic field. The third step consists of periodic drift calibration during measurement. The results and detailed description of each calibration step are presented and discussed in the paper. The gradiometer is finally verified during the detection of the metal object in the measuring grid

  4. ORNL calibrations facility

    International Nuclear Information System (INIS)

    Berger, C.D.; Gupton, E.D.; Lane, B.H.; Miller, J.H.; Nichols, S.W.

    1982-08-01

    The ORNL Calibrations Facility is operated by the Instrumentation Group of the Industrial Safety and Applied Health Physics Division. Its primary purpose is to maintain radiation calibration standards for calibration of ORNL health physics instruments and personnel dosimeters. This report includes a discussion of the radioactive sources and ancillary equipment in use and a step-by-step procedure for calibration of those survey instruments and personnel dosimeters in routine use at ORNL

  5. Calibration procedure for thermoluminescent dosemeters in water absorbed doses for Iridium-192 high dose rate sources

    International Nuclear Information System (INIS)

    Reyes Cac, Franky Eduardo

    2004-10-01

    Thermoluminescent dosimeters are used in brachytherapy services quality assurance programs, with the aim of guaranteeing the correct radiation dose supplied to cancer patients, as well as with the purpose of evaluating new clinical procedures. This work describes a methodology for thermoluminescent dosimeters calibration in terms of absorbed dose to water for 192 Ir high dose rate sources. The reference dose used is measured with an ionization chamber previously calibrated for 192 Ir energy quality, applying the methodology proposed by Toelli. This methodology aims to standardizing the procedure, in a similar form to that used for external radiotherapy. The work evolves the adaptation of the TRS-277 Code of the International Atomic Energy Agency, for small and big cavities, through the introduction for non-uniform experimental factor, for the absorbed dose in the neighborhood of small brachytherapy sources. In order to simulate a water medium around the source during the experimental work, an acrylic phantom was used. It guarantees the reproducibility of the ionization chamber and the thermoluminescent dosimeter's location in relation to the radiation source. The values obtained with the ionization chamber and the thermoluminescent dosimeters, exposed to a 192 Ir high dose rate source, were compared and correction factors for different source-detector distances were determined for the thermoluminescent dosimeters. A numeric function was generated relating the correction factors and the source-detector distance. These correction factors are in fact the thermoluminescent dosimeter calibration factors for the 192 Ir source considered. As a possible application of this calibration methodology for thermoluminescent dosimeters, a practical range of source-detector distances is proposed for quality control of 192 Ir high dose rate sources. (author)

  6. Calibrating nacelle lidars

    Energy Technology Data Exchange (ETDEWEB)

    Courtney, M.

    2013-01-15

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  7. Observatory Magnetometer In-Situ Calibration

    Directory of Open Access Journals (Sweden)

    A Marusenkov

    2011-07-01

    Full Text Available An experimental validation of the in-situ calibration procedure, which allows estimating parameters of observatory magnetometers (scale factors, sensor misalignment without its operation interruption, is presented. In order to control the validity of the procedure, the records provided by two magnetometers calibrated independently in a coil system have been processed. The in-situ estimations of the parameters are in very good agreement with the values provided by the coil system calibration.

  8. Calibration Of Partial-Pressure-Of-Oxygen Sensors

    Science.gov (United States)

    Yount, David W.; Heronimus, Kevin

    1995-01-01

    Report and analysis of, and discussion of improvements in, procedure for calibrating partial-pressure-of-oxygen sensors to satisfy Spacelab calibration requirements released. Sensors exhibit fast drift, which results in short calibration period not suitable for Spacelab. By assessing complete process of determining total drift range available, calibration procedure modified to eliminate errors and still satisfy requirements without compromising integrity of system.

  9. Thermal power calibrations of the IPR-R1 TRIGA reactor by the calorimetric and the heat balance methods

    International Nuclear Information System (INIS)

    Mesquita, Amir Zacarias; Rezende, Hugo Cesar; Souza, Rose Mary Gomes do Prado

    2009-01-01

    Since the first nuclear reactor was built, a number of methodological variations have been evolved for the calibration of the reactor thermal power. Power monitoring of reactors is done by means of neutronic instruments, but its calibration is always done by thermal procedures. The purpose of this paper is to present the results of the thermal power calibration carried out on March 5th, 2009 in the IPR-R1 TRIGA reactor. It was used two procedures: the calorimetric and heat balance methods. The calorimetric procedure was done with the reactor operating at a constant power, with primary cooling system switched off. The rate of temperature rise of the water was recorded. The reactor power is calculate as a function of the temperature-rise rate and the system heat capacity constant. The heat balance procedure consists in the steady-state energy balance of the primary cooling loop of the reactor. For this balance, the inlet and outlet temperatures and the water flow in the primary cooling loop were measured. The heat transferred through the primary loop was added to the heat leakage from the reactor pool. The calorimetric method calibration presented a large uncertainty. The main source of error was the determination of the heat content of the system, due to a large uncertainty in the volume of the water in the system and a lack of homogenization of the water temperature. The heat balance calibration in the primary loop is the standard procedure for calibrating the power of the IPR-R1 TRIGA nuclear reactor. (author))

  10. Cryogenic Pressure Calibration Facility Using a Cold Force Reference

    CERN Document Server

    Bager, T; Métral, L

    1999-01-01

    Presently various commercial cryogenic pressure sensors are being investigated for installation in the LHC collider, they will eventually be used to assess that the magnets are fully immersed in liquid and to monitor fast pressure transients. In the framework of this selection procedure a cryogenic pressue calibration facility has been designed and built; it is based on a cryogenic primary pressure reference made of a bellows that converts the pressure into a force measurement. For that a shaft transfers this force to a precision force transducer at room temperature. Knowing the liquid bath pessure and the surface area of the bellows the pressure applied to the transducers under calibration is calculated; corrections due to thermal contraction are introduced. To avoid loss of force in the bellows wall its length is maintained constant; a cold capacitive displacement sensor measures this. The calibration temperature covers 1.5 K to 4.2 K and the pressure 0 to 20 bar. In contrast with more classical techniques ...

  11. Calibration procedures for mammography dosemeters in Poland

    International Nuclear Information System (INIS)

    Gwiazdowska, B.; Ulkowski, P.; Tolwinski, J.; Bulski, W.

    2002-01-01

    Breast cancer is the most frequent tumour in women and the effectiveness of the treatment depends dramatically on the early detection of the disease. That is the reason why in Poland the mammography control examinations are strongly supported by the Centre of Oncology. In Poland there are over 400 mammography units which account for about 300,000 examinations per year. An investigation performed by the Medical Physics Department of the Centre of Oncology in Warsaw at about 100 mammography facilities proved that in most cases the doses absorbed by the patients could be reduced without decrease of image quality. This is one of the reasons why the Polish Secondary Standard Dosimetry Laboratory (SSDL) dealing mainly with calibration of radiotherapy dosemeters is extending its activities and therefore new facilities and equipment adapted for calibration of mammographic dosemeters have been installed. The mammography dosimetry calibration equipment is permanently installed in the same laboratory room where the radiotherapy dosemeters are calibrated. A base of a mammography unit no longer in clinical use, together with its movable system has been adapted to handle ionization chamber holders. An X-ray tube with a 50 kV high frequency generator was also installed. The tube, a Varian type OEG-50-2, (designed for laboratory applications) with molybdenum anode of an anode angle 23,7 deg. and with a large focus, effective size approximately 5 mm 2 , has an inherent filtration of 0,25 mm beryllium. It is installed in a housing with 2mm lead shielding; a cone shaped beam is formed by a system of three collimators

  12. Validation of an efficiency calibration procedure for a coaxial n-type and a well-type HPGe detector used for the measurement of environmental radioactivity

    Energy Technology Data Exchange (ETDEWEB)

    Morera-Gómez, Yasser, E-mail: ymore24@gamail.com [Centro de Estudios Ambientales de Cienfuegos, AP 5. Ciudad Nuclear, CP 59350 Cienfuegos (Cuba); Departamento de Química y Edafología, Universidad de Navarra, Irunlarrea No 1, Pamplona 31009, Navarra (Spain); Cartas-Aguila, Héctor A.; Alonso-Hernández, Carlos M.; Nuñez-Duartes, Carlos [Centro de Estudios Ambientales de Cienfuegos, AP 5. Ciudad Nuclear, CP 59350 Cienfuegos (Cuba)

    2016-05-11

    To obtain reliable measurements of the environmental radionuclide activity using HPGe (High Purity Germanium) detectors, the knowledge of the absolute peak efficiency is required. This work presents a practical procedure for efficiency calibration of a coaxial n-type and a well-type HPGe detector using experimental and Monte Carlo simulations methods. The method was performed in an energy range from 40 to 1460 keV and it can be used for both, solid and liquid environmental samples. The calibration was initially verified measuring several reference materials provided by the IAEA (International Atomic Energy Agency). Finally, through the participation in two Proficiency Tests organized by IAEA for the members of the ALMERA network (Analytical Laboratories for the Measurement of Environmental Radioactivity) the validity of the developed procedure was confirmed. The validation also showed that measurement of {sup 226}Ra should be conducted using coaxial n-type HPGe detector in order to minimize the true coincidence summing effect. - Highlights: • An efficiency calibration for a coaxial and a well-type HPGe detector was performed. • The calibration was made using experimental and Monte Carlo simulations methods. • The procedure was verified measuring several reference materials provided by IAEA. • Calibrations were validated through the participation in 2 ALMERA Proficiency Tests.

  13. Problems involved in quantitative gamma camera scintigraphy. A. Procedure and definitions

    International Nuclear Information System (INIS)

    Erbsmann, F.

    1976-01-01

    In the procedure used the elementary notions on calibration sources (isotropic point source, parallel-beam point source, tubular line source, parallel-ray line source, sources of finite dimensions, homogeneous phantom) are reviewed, together with the most frequently used definitions, those of the scattering function and of the modulation transfer functions [fr

  14. Calibration of thermoluminiscent materials

    International Nuclear Information System (INIS)

    Bos, A.J.J.

    1989-07-01

    In this report the relation between exposure and absorbed radiation dose in various materials is represented, on the base of recent data. With the help of this a calibration procedure for thermoluminescent materials, adapted to the IRI radiation standard is still the exposure in rontgen. In switching to the air kerma standard the calibration procedure will have to be adapted. (author). 6 refs.; 4 tabs

  15. Calibration Procedures in Mid Format Camera Setups

    Science.gov (United States)

    Pivnicka, F.; Kemper, G.; Geissler, S.

    2012-07-01

    A growing number of mid-format cameras are used for aerial surveying projects. To achieve a reliable and geometrically precise result also in the photogrammetric workflow, awareness on the sensitive parts is important. The use of direct referencing systems (GPS/IMU), the mounting on a stabilizing camera platform and the specific values of the mid format camera make a professional setup with various calibration and misalignment operations necessary. An important part is to have a proper camera calibration. Using aerial images over a well designed test field with 3D structures and/or different flight altitudes enable the determination of calibration values in Bingo software. It will be demonstrated how such a calibration can be performed. The direct referencing device must be mounted in a solid and reliable way to the camera. Beside the mechanical work especially in mounting the camera beside the IMU, 2 lever arms have to be measured in mm accuracy. Important are the lever arms from the GPS Antenna to the IMU's calibrated centre and also the lever arm from the IMU centre to the Camera projection centre. In fact, the measurement with a total station is not a difficult task but the definition of the right centres and the need for using rotation matrices can cause serious accuracy problems. The benefit of small and medium format cameras is that also smaller aircrafts can be used. Like that, a gyro bases stabilized platform is recommended. This causes, that the IMU must be mounted beside the camera on the stabilizer. The advantage is, that the IMU can be used to control the platform, the problematic thing is, that the IMU to GPS antenna lever arm is floating. In fact we have to deal with an additional data stream, the values of the movement of the stabiliser to correct the floating lever arm distances. If the post-processing of the GPS-IMU data by taking the floating levers into account, delivers an expected result, the lever arms between IMU and camera can be applied

  16. Calibration-on-the-spot”: How to calibrate an EMCCD camera from its images

    DEFF Research Database (Denmark)

    Mortensen, Kim; Flyvbjerg, Henrik

    2016-01-01

    In order to count photons with a camera, the camera must be calibrated. Photon counting is necessary, e.g., to determine the precision of localization-based super-resolution microscopy. Here we present a protocol that calibrates an EMCCD camera from information contained in isolated, diffraction-......-limited spots in any image taken by the camera, thus making dedicated calibration procedures redundant by enabling calibration post festum, from images filed without calibration information....

  17. Calibration of nuclear medicine gamma counters

    International Nuclear Information System (INIS)

    Orlic, M.; Spasic-Jokic, V.; Jovanovic, M.; Vranjes, S. . E-mail address of corresponding author: morlic@vin.bg.ac.yu; Orlic, M.)

    2005-01-01

    In this paper the practical problem of nuclear medicine gamma counters calibration has been solved by using dose calibrators CRC-15R with standard error ±5%. The samples from technetium generators have been measured both by dose calibrators CRC-15R and gamma counter ICN Gamma 3.33 taking into account decay correction. Only the linear part of the curve has practical meaning. The advantage of this procedure satisfies the requirements from international standards: the calibration of sources used for medical exposure be traceable to a standard dosimetry laboratory and radiopharmaceuticals for nuclear medicine procedures be calibrated in terms of activity of the radiopharmaceutical to be administered. (author)

  18. Calibrating the neutron moisture meter: Precision and economy

    International Nuclear Information System (INIS)

    Akhter, J.; Waheed, R.A.; Hignett, C.T.; Greacen, E.L.

    2000-01-01

    Established laboratory and field calibration procedures for the neutron moisture meter are demonstrated on a uniform soil and alternative, low cost procedures on a duplex, less uniform soil. The effect of field variability on the calibration methodology is discussed with the aim of optimising calibration reliability at minimal cost. The difference between calibration for a soil material, or for a field (a range of soil materials) is considered. In particular, calibration for the estimation of water content change is shown to be a different problem from calibration for the estimation of water content in a variable field. Techniques aimed at detecting field variability problems during calibration are suggested, and methods for optimising the results for the intended use of the instrument are outlined. Pairing of calibration tubes, alternative methods of analysis of calibration data, and use of other information from the field to measure its variability, can improve the precision of calibration procedures to the point where minimal calibration effort, with careful analysis, can provide reliable results. (author)

  19. A novel single-step procedure for the calibration of the mounting parameters of a multi-camera terrestrial mobile mapping system

    Science.gov (United States)

    Habib, A.; Kersting, P.; Bang, K.; Rau, J.

    2011-12-01

    Mobile Mapping Systems (MMS) can be defined as moving platforms which integrates a set of imaging sensors and a position and orientation system (POS) for the collection of geo-spatial information. In order to fully explore the potential accuracy of such systems and guarantee accurate multi-sensor integration, a careful system calibration must be carried out. System calibration involves individual sensor calibration as well as the estimation of the inter-sensor geometric relationship. This paper tackles a specific component of the system calibration process of a multi-camera MMS - the estimation of the relative orientation parameters among the cameras, i.e., the inter-camera geometric relationship (lever-arm offsets and boresight angles among the cameras). For that purpose, a novel single step procedure, which is easy to implement and not computationally intensive, will be introduced. The proposed method is implemented in such a way that it can also be used for the estimation of the mounting parameters among the cameras and the IMU body frame, in case of directly georeferenced systems. The performance of the proposed method is evaluated through experimental results using simulated data. A comparative analysis between the proposed single-step and the two-step, which makes use of the traditional bundle adjustment procedure, is demonstrated.

  20. SPOTS Calibration Example

    Directory of Open Access Journals (Sweden)

    Patterson E.

    2010-06-01

    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  1. CALIBRATION PROCEDURES IN MID FORMAT CAMERA SETUPS

    Directory of Open Access Journals (Sweden)

    F. Pivnicka

    2012-07-01

    Full Text Available A growing number of mid-format cameras are used for aerial surveying projects. To achieve a reliable and geometrically precise result also in the photogrammetric workflow, awareness on the sensitive parts is important. The use of direct referencing systems (GPS/IMU, the mounting on a stabilizing camera platform and the specific values of the mid format camera make a professional setup with various calibration and misalignment operations necessary. An important part is to have a proper camera calibration. Using aerial images over a well designed test field with 3D structures and/or different flight altitudes enable the determination of calibration values in Bingo software. It will be demonstrated how such a calibration can be performed. The direct referencing device must be mounted in a solid and reliable way to the camera. Beside the mechanical work especially in mounting the camera beside the IMU, 2 lever arms have to be measured in mm accuracy. Important are the lever arms from the GPS Antenna to the IMU's calibrated centre and also the lever arm from the IMU centre to the Camera projection centre. In fact, the measurement with a total station is not a difficult task but the definition of the right centres and the need for using rotation matrices can cause serious accuracy problems. The benefit of small and medium format cameras is that also smaller aircrafts can be used. Like that, a gyro bases stabilized platform is recommended. This causes, that the IMU must be mounted beside the camera on the stabilizer. The advantage is, that the IMU can be used to control the platform, the problematic thing is, that the IMU to GPS antenna lever arm is floating. In fact we have to deal with an additional data stream, the values of the movement of the stabiliser to correct the floating lever arm distances. If the post-processing of the GPS-IMU data by taking the floating levers into account, delivers an expected result, the lever arms between IMU and

  2. Bayesian calibration of coarse-grained forces: Efficiently addressing transferability

    International Nuclear Information System (INIS)

    Patrone, Paul N.; Rosch, Thomas W.; Phelan, Frederick R.

    2016-01-01

    Generating and calibrating forces that are transferable across a range of state-points remains a challenging task in coarse-grained (CG) molecular dynamics. In this work, we present a coarse-graining workflow, inspired by ideas from uncertainty quantification and numerical analysis, to address this problem. The key idea behind our approach is to introduce a Bayesian correction algorithm that uses functional derivatives of CG simulations to rapidly and inexpensively recalibrate initial estimates f 0 of forces anchored by standard methods such as force-matching. Taking density-temperature relationships as a running example, we demonstrate that this algorithm, in concert with various interpolation schemes, can be used to efficiently compute physically reasonable force curves on a fine grid of state-points. Importantly, we show that our workflow is robust to several choices available to the modeler, including the interpolation schemes and tools used to construct f 0 . In a related vein, we also demonstrate that our approach can speed up coarse-graining by reducing the number of atomistic simulations needed as inputs to standard methods for generating CG forces.

  3. Bayesian calibration of coarse-grained forces: Efficiently addressing transferability

    Science.gov (United States)

    Patrone, Paul N.; Rosch, Thomas W.; Phelan, Frederick R.

    2016-04-01

    Generating and calibrating forces that are transferable across a range of state-points remains a challenging task in coarse-grained (CG) molecular dynamics. In this work, we present a coarse-graining workflow, inspired by ideas from uncertainty quantification and numerical analysis, to address this problem. The key idea behind our approach is to introduce a Bayesian correction algorithm that uses functional derivatives of CG simulations to rapidly and inexpensively recalibrate initial estimates f0 of forces anchored by standard methods such as force-matching. Taking density-temperature relationships as a running example, we demonstrate that this algorithm, in concert with various interpolation schemes, can be used to efficiently compute physically reasonable force curves on a fine grid of state-points. Importantly, we show that our workflow is robust to several choices available to the modeler, including the interpolation schemes and tools used to construct f0. In a related vein, we also demonstrate that our approach can speed up coarse-graining by reducing the number of atomistic simulations needed as inputs to standard methods for generating CG forces.

  4. Linear Calibration – Is It so Simple?

    International Nuclear Information System (INIS)

    Arsova, Diana; Babanova, Sofia; Mandjukov, Petko

    2009-01-01

    Calibration procedure is an important part of instrumental analysis. Usually it is not the major uncertainty source in whole analytical procedure. However, improper calibration might cause a significant bias of the analytical results from the real (certified) value. Standard Gaussian linear regression is the most frequently used mathematical approach for estimation of calibration function parameters. In the present article are discussed some not quite popular, but highly recommended in certain cases methods for parameter estimation, such as: weighted regression, orthogonal regression, robust regression, bracketing calibration etc. Some useful approximations are also presented. Special attention is paid to the statistical criteria which to be used for selection of proper calibration model. Standard UV-VIS spectrometric procedure for determination of phosphates in water was used as a practical example. Several different approaches for estimation of the contribution of calibration to the general un-certainty of the analytical result are presented and compared

  5. EM simulation assisted parameter extraction for the modeling of transferred-substrate InP HBTs

    DEFF Research Database (Denmark)

    Johansen, Tom Keinicke; Weimann, Nils; Doerner, Ralf

    2017-01-01

    In this paper an electromagnetic (EM) simulation assisted parameters extraction procedure is demonstrated for accurate modeling of down-scaled transferred-substrate InP HBTs. The external parasitic network associated with via transitions and device electrodes is carefully extracted from calibrate...

  6. Mathematical Model and Calibration Procedure of a PSD Sensor Used in Local Positioning Systems.

    Science.gov (United States)

    Rodríguez-Navarro, David; Lázaro-Galilea, José Luis; Bravo-Muñoz, Ignacio; Gardel-Vicente, Alfredo; Domingo-Perez, Francisco; Tsirigotis, Georgios

    2016-09-15

    Here, we propose a mathematical model and a calibration procedure for a PSD (position sensitive device) sensor equipped with an optical system, to enable accurate measurement of the angle of arrival of one or more beams of light emitted by infrared (IR) transmitters located at distances of between 4 and 6 m. To achieve this objective, it was necessary to characterize the intrinsic parameters that model the system and obtain their values. This first approach was based on a pin-hole model, to which system nonlinearities were added, and this was used to model the points obtained with the nA currents provided by the PSD. In addition, we analyzed the main sources of error, including PSD sensor signal noise, gain factor imbalances and PSD sensor distortion. The results indicated that the proposed model and method provided satisfactory calibration and yielded precise parameter values, enabling accurate measurement of the angle of arrival with a low degree of error, as evidenced by the experimental results.

  7. Calibration and energy response of the Bitt RM10/RS02 gamma radiation detectors

    International Nuclear Information System (INIS)

    Dijk, E. van; Aalbers, A.H.L.

    1990-03-01

    A radiation monitoring network with automatic warning capabilities (LMR) has been established in the Netherlands. For the detection of gamma radiation exposure-rate-meters manufactured by Bitt Technologies are used. These meters consist of a proportional counter tube (type RS 02) and a read-out unit (type RM 10E). The photon energy response of 6 counter tubes was tested at the National Institute of Public Health and Environmental Protection. The measurements were performed with heavy filtered X-rays in the range of 50-250 keV (ISO narrow spectrum series) and with gamma ray beams from cesium-137 (662 keV) and cobalt-60 (1,25 MeV). To determine the energy response, the detector reading was referred to air kerma by means of a transfer ionization chamber. This transfer chamber was directly calibrated against the standard for X-rays. By applying these measurement procedures of a set of calibration factors (N k ) as a function of photon energy was determined. These calibration factors, expressed as the ratio air kerma to reading were converted to ambient dose equivalent calibration factors using appropriate conversion factors taken from Grosswend et al., 1988. From the measurement data an average ambient dose equivalent calibration factor of 10.8 mSv.roentgen -1 was calculated. (author). 5 refs.; 6 figs.; 5 tabs

  8. Intercomparison and calibration of dose calibrators used in nuclear medicine facilities

    CERN Document Server

    Costa, A M D

    2003-01-01

    The aim of this work was to establish a working standard for intercomparison and calibration of dose calibrators used in most of nuclear medicine facilities for the determination of the activity of radionuclides administered to patients in specific examinations or therapeutic procedures. A commercial dose calibrator, a set of standard radioactive sources, and syringes, vials and ampoules with radionuclide solutions used in nuclear medicine were utilized in this work. The commercial dose calibrator was calibrated for radionuclide solutions used in nuclear medicine. Simple instrument tests, such as linearity response and variation response with the source volume at a constant source activity concentration were performed. This instrument may be used as a reference system for intercomparison and calibration of other activity meters, as a method of quality control of dose calibrators utilized in nuclear medicine facilities.

  9. Exposure-rate calibration using large-area calibration pads

    International Nuclear Information System (INIS)

    Novak, E.F.

    1988-09-01

    The US Department of Energy (DOE) Office of Remedial Action and Waste Technology established the Technical Measurements Center (TMC) at the DOE Grand Junction Projects Office (GJPO) in Grand Junction, Colorado, to standardize, calibrate, and compare measurements made in support of DOE remedial action programs. A set of large-area, radioelement-enriched concrete pads was constructed by the DOE in 1978 at the Walker Field Airport in Grand Junction for use as calibration standards for airborne gamma-ray spectrometer systems. The use of these pads was investigated by the TMC as potential calibration standards for portable scintillometers employed in measuring gamma-ray exposure rates at Uranium Mill Tailings Remedial Action (UMTRA) project sites. Data acquired on the pads using a pressurized ionization chamber (PIC) and three scintillometers are presented as an illustration of an instrumental calibration. Conclusions and recommended calibration procedures are discussed, based on the results of these data

  10. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    International Nuclear Information System (INIS)

    Cornic, Philippe; Le Besnerais, Guy; Champagnat, Frédéric; Illoul, Cédric; Cheminet, Adam; Le Sant, Yves; Leclaire, Benjamin

    2016-01-01

    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data. (paper)

  11. Lidar to lidar calibration phase 2

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents the results from phase 2 of a lidar to lidar (L2L) calibration procedure. Phase two of the project included two measurement campaigns conducted at given sites. The purpose was to find out if the lidar-to-lidar calibration procedure can be conducted with similar results...

  12. 40 CFR 90.424 - Dilute sampling procedures-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... calibration. 90.424 Section 90.424 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR... “true” value) for the CVS calibration. (Note: In no case should an upstream screen or other restriction... Parameter Symbol Units Sensor-readout tolerances Barometric pressure (corrected) PB kPa ±.340 kPa. Ambient...

  13. A finite volume procedure for fluid flow, heat transfer and solid-body stress analysis

    KAUST Repository

    Jagad, P. I.; Puranik, B. P.; Date, A. W.

    2018-01-01

    A unified cell-centered unstructured mesh finite volume procedure is presented for fluid flow, heat transfer and solid-body stress analysis. An in-house procedure (A. W. Date, Solution of Transport Equations on Unstructured Meshes with Cell

  14. A finite volume procedure for fluid flow, heat transfer and solid-body stress analysis

    KAUST Repository

    Jagad, P. I.

    2018-04-12

    A unified cell-centered unstructured mesh finite volume procedure is presented for fluid flow, heat transfer and solid-body stress analysis. An in-house procedure (A. W. Date, Solution of Transport Equations on Unstructured Meshes with Cell-Centered Colocated Variables. Part I: Discretization, International Journal of Heat and Mass Transfer, vol. 48 (6), 1117-1127, 2005) is extended to include the solid-body stress analysis. The transport terms for a cell-face are evaluated in a structured grid-like manner. The Cartesian gradients at the center of each cell-face are evaluated using the coordinate transformation relations. The accuracy of the procedure is demonstrated by solving several benchmark problems involving different boundary conditions, source terms, and types of loading.

  15. Mechanics of log calibration

    International Nuclear Information System (INIS)

    Waller, W.C.; Cram, M.E.; Hall, J.E.

    1975-01-01

    For any measurement to have meaning, it must be related to generally accepted standard units by a valid and specified system of comparison. To calibrate well-logging tools, sensing systems are designed which produce consistent and repeatable indications over the range for which the tool was intended. The basics of calibration theory, procedures, and calibration record presentations are reviewed. Calibrations for induction, electrical, radioactivity, and sonic logging tools will be discussed. The authors' intent is to provide an understanding of the sources of errors, of the way errors are minimized in the calibration process, and of the significance of changes in recorded calibration data

  16. Tablet potency of Tianeptine in coated tablets by near infrared spectroscopy: model optimisation, calibration transfer and confidence intervals.

    Science.gov (United States)

    Boiret, Mathieu; Meunier, Loïc; Ginot, Yves-Michel

    2011-02-20

    A near infrared (NIR) method was developed for determination of tablet potency of active pharmaceutical ingredient (API) in a complex coated tablet matrix. The calibration set contained samples from laboratory and production scale batches. The reference values were obtained by high performance liquid chromatography (HPLC) and partial least squares (PLS) regression was used to establish a model. The model was challenged by calculating tablet potency of two external test sets. Root mean square errors of prediction were respectively equal to 2.0% and 2.7%. To use this model with a second spectrometer from the production field, a calibration transfer method called piecewise direct standardisation (PDS) was used. After the transfer, the root mean square error of prediction of the first test set was 2.4% compared to 4.0% without transferring the spectra. A statistical technique using bootstrap of PLS residuals was used to estimate confidence intervals of tablet potency calculations. This method requires an optimised PLS model, selection of the bootstrap number and determination of the risk. In the case of a chemical analysis, the tablet potency value will be included within the confidence interval calculated by the bootstrap method. An easy to use graphical interface was developed to easily determine if the predictions, surrounded by minimum and maximum values, are within the specifications defined by the regulatory organisation. Copyright © 2010 Elsevier B.V. All rights reserved.

  17. Test of an albedo neutron dosimetry system: TLD calibration and readout procedure, neutron calibration, dosimetry properties, routine application

    International Nuclear Information System (INIS)

    Piesch, E.; Burgkhardt, B.

    1988-03-01

    The two-component albedo dosemeter in use consists of an universal boron-loaded plastic encapsulation, the beta and albedo neutron windows of which are adopted to the corresponding TLD system of the manufacturers Alnor, Harshaw, Panasonic and Vinten. Beside the TLD detectors the capsule may contain also track etch detectors. Within a BMU project the system was investigated by four governmental measurement services in the FRG with respect to its qualification for personnel monitoring with emphasis in the readout and calibration procedures for the TLD system, the evaluation technique for the estimation of the photon and neutron dose equivalent in routine monitoring and the calibration of the personnel dosemeter in stray neutron fields. The test has shown the readiness of the system to act in the application areas of nuclear power reactors and linacs behind heavy shieldings, in the fuel element cycle, use of fissile materials, criticality, use of radionuclide sources, high energy particle accelerators. The uncertainty due to energy dependence was found to be within a factor of 2 for a single application area. In the case of irradiations from the front half space the dose equivalent H'(10) is indicated sufficiently independent of the direction of the radiation incidence. After completion of the test the albedo dosemeter became the official neutron personnel dosemeter in the FRG. It allows the separate estimation of the dose equivalent of hard beta radiation, photon radiation and neutrons. (orig./HP) [de

  18. REMINDER Saved Leave Scheme (SLS) : Simplified procedure for the transfer of leave to saved leave accounts

    CERN Multimedia

    HR Division

    2001-01-01

    As part of the process of streamlining procedures, the HR and AS Divisions have jointly developed a system whereby annual and compensatory leave will henceforth be automatically transferred1) to saved leave accounts. Under the provisions of the voluntary saved leave scheme (SLS), a maximum total of 10 days'2)Previously, every person taking part in the scheme has been individually issued with a form for the purposes of requesting the transfer of leave to the leave account and the transfer has then had to be done manually by HR Division. To streamline the procedure, unused leave of all those taking part in the saved leave scheme at the closure of the leave-year accounts will henceforth be transferred automatically to the saved leave account on that date. This simplification is in the interest of all parties concerned. This automatic transfer procedure has a number of advantages for participants in the SLS scheme. First, staff members will no longer have to take any administrative steps. Secondly, the new proced...

  19. Calibration procedure of Hukseflux SR25 to Establish the Diffuse Reference for the Outdoor Broadband Radiometer Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Reda, Ibrahim M. [National Renewable Energy Laboratory (NREL), Golden, CO (United States); Andreas, Afshin M. [National Renewable Energy Laboratory (NREL), Golden, CO (United States)

    2017-08-01

    Accurate pyranometer calibrations, traceable to internationally recognized standards, are critical for solar irradiance measurements. One calibration method is the component summation method, where the pyranometers are calibrated outdoors under clear sky conditions, and the reference global solar irradiance is calculated as the sum of two reference components, the diffuse horizontal and subtended beam solar irradiances. The beam component is measured with pyrheliometers traceable to the World Radiometric Reference, while there is no internationally recognized reference for the diffuse component. In the absence of such a reference, we present a method to consistently calibrate pyranometers for measuring the diffuse component. The method is based on using a modified shade/unshade method and a pyranometer with less than 0.5 W/m2 thermal offset. The calibration result shows that the responsivity of Hukseflux SR25 pyranometer equals 10.98 uV/(W/m2) with +/-0.86 percent uncertainty.

  20. Nitrous oxide emissions from cropland: a procedure for calibrating the DayCent biogeochemical model using inverse modelling

    Science.gov (United States)

    Rafique, Rashad; Fienen, Michael N.; Parkin, Timothy B.; Anex, Robert P.

    2013-01-01

    DayCent is a biogeochemical model of intermediate complexity widely used to simulate greenhouse gases (GHG), soil organic carbon and nutrients in crop, grassland, forest and savannah ecosystems. Although this model has been applied to a wide range of ecosystems, it is still typically parameterized through a traditional “trial and error” approach and has not been calibrated using statistical inverse modelling (i.e. algorithmic parameter estimation). The aim of this study is to establish and demonstrate a procedure for calibration of DayCent to improve estimation of GHG emissions. We coupled DayCent with the parameter estimation (PEST) software for inverse modelling. The PEST software can be used for calibration through regularized inversion as well as model sensitivity and uncertainty analysis. The DayCent model was analysed and calibrated using N2O flux data collected over 2 years at the Iowa State University Agronomy and Agricultural Engineering Research Farms, Boone, IA. Crop year 2003 data were used for model calibration and 2004 data were used for validation. The optimization of DayCent model parameters using PEST significantly reduced model residuals relative to the default DayCent parameter values. Parameter estimation improved the model performance by reducing the sum of weighted squared residual difference between measured and modelled outputs by up to 67 %. For the calibration period, simulation with the default model parameter values underestimated mean daily N2O flux by 98 %. After parameter estimation, the model underestimated the mean daily fluxes by 35 %. During the validation period, the calibrated model reduced sum of weighted squared residuals by 20 % relative to the default simulation. Sensitivity analysis performed provides important insights into the model structure providing guidance for model improvement.

  1. Restoring tactile and erogenous penile sensation in low-spinal-lesion patients: procedural and technical aspects following 43 TOMAX nerve transfer procedures.

    Science.gov (United States)

    Overgoor, Max L E; de Jong, Tom P V M; Kon, Moshe

    2014-08-01

    The "TOMAX" (TO MAX-imize sensation, sexuality, and quality of life) procedure restores genital sensation in men with low spinal lesions, improving sexual health, as shown previously. It connects the dorsal nerve of the penis to the intact ipsilateral ilioinguinal nerve, unilaterally or bilaterally. This study reports on the technical aspects based on 43 TOMAX nerve transfers. In 40 patients with no penile but intact groin sensation, 43 nerve transfers were performed. Data on patient selection, surgical history, anatomy of the ilioinguinal nerve and dorsal nerve of the penis, unilateral or bilateral surgery, surgical technique, complications, and patient information were collected prospectively. Regardless of origin, all patients with no penile but good groin sensation are eligible for the procedure, provided the ilioinguinal nerve is not damaged because of former inguinal surgery or absent because of anatomical variations. Selection of a unilateral or bilateral procedure depends on the presence or absence of reflex erections and bulbocavernosus reflex. Preliminary experience with the first three bilateral cases shows that it is technically feasible, with encouraging results. The surgical technique has evolved (described in detail, including video) to enhance outcome and reduce complications. Patients are better informed, resulting in realistic expectations. This article synthesizes the procedural and technical experience of 43 TOMAX nerve transfers. Anyone skilled in peripheral nerve surgery and microsurgery can adopt this concept and further develop it. The TOMAX procedure can then be used to restore erogenous penile sensation and improve the quality of sexual health in patients with absent penile but good groin sensation.

  2. Reconstructing the calibrated strain signal in the Advanced LIGO detectors

    Science.gov (United States)

    Viets, A. D.; Wade, M.; Urban, A. L.; Kandhasamy, S.; Betzwieser, J.; Brown, Duncan A.; Burguet-Castell, J.; Cahillane, C.; Goetz, E.; Izumi, K.; Karki, S.; Kissel, J. S.; Mendell, G.; Savage, R. L.; Siemens, X.; Tuyenbayev, D.; Weinstein, A. J.

    2018-05-01

    Advanced LIGO’s raw detector output needs to be calibrated to compute dimensionless strain h(t) . Calibrated strain data is produced in the time domain using both a low-latency, online procedure and a high-latency, offline procedure. The low-latency h(t) data stream is produced in two stages, the first of which is performed on the same computers that operate the detector’s feedback control system. This stage, referred to as the front-end calibration, uses infinite impulse response (IIR) filtering and performs all operations at a 16 384 Hz digital sampling rate. Due to several limitations, this procedure currently introduces certain systematic errors in the calibrated strain data, motivating the second stage of the low-latency procedure, known as the low-latency gstlal calibration pipeline. The gstlal calibration pipeline uses finite impulse response (FIR) filtering to apply corrections to the output of the front-end calibration. It applies time-dependent correction factors to the sensing and actuation components of the calibrated strain to reduce systematic errors. The gstlal calibration pipeline is also used in high latency to recalibrate the data, which is necessary due mainly to online dropouts in the calibrated data and identified improvements to the calibration models or filters.

  3. Normalization and calibration of geostationary satellite radiances for the International Satellite Cloud Climatology Project

    Science.gov (United States)

    Desormeaux, Yves; Rossow, William B.; Brest, Christopher L.; Campbell, G. G.

    1993-01-01

    Procedures are described for normalizing the radiometric calibration of image radiances obtained from geostationary weather satellites that contributed data to the International Satellite Cloud Climatology Project. The key step is comparison of coincident and collocated measurements made by each satellite and the concurrent AVHRR on the 'afternoon' NOAA polar-orbiting weather satellite at the same viewing geometry. The results of this comparison allow transfer of the AVHRR absolute calibration, which has been established over the whole series, to the radiometers on the geostationary satellites. Results are given for Meteosat-2, 3, and 4, for GOES-5, 6, and 7, for GMS-2, 3, and 4 and for Insat-1B. The relative stability of the calibrations of these radiance data is estimated to be within +/- 3 percent; the uncertainty of the absolute calibrations is estimated to be less than 10 percent. The remaining uncertainties are at least two times smaller than for the original radiance data.

  4. Line fiducial material and thickness considerations for ultrasound calibration

    Science.gov (United States)

    Ameri, Golafsoun; McLeod, A. J.; Baxter, John S. H.; Chen, Elvis C. S.; Peters, Terry M.

    2015-03-01

    Ultrasound calibration is a necessary procedure in many image-guided interventions, relating the position of tools and anatomical structures in the ultrasound image to a common coordinate system. This is a necessary component of augmented reality environments in image-guided interventions as it allows for a 3D visualization where other surgical tools outside the imaging plane can be found. Accuracy of ultrasound calibration fundamentally affects the total accuracy of this interventional guidance system. Many ultrasound calibration procedures have been proposed based on a variety of phantom materials and geometries. These differences lead to differences in representation of the phantom on the ultrasound image which subsequently affect the ability to accurately and automatically segment the phantom. For example, taut wires are commonly used as line fiducials in ultrasound calibration. However, at large depths or oblique angles, the fiducials appear blurred and smeared in ultrasound images making it hard to localize their cross-section with the ultrasound image plane. Intuitively, larger diameter phantoms with lower echogenicity are more accurately segmented in ultrasound images in comparison to highly reflective thin phantoms. In this work, an evaluation of a variety of calibration phantoms with different geometrical and material properties for the phantomless calibration procedure was performed. The phantoms used in this study include braided wire, plastic straws, and polyvinyl alcohol cryogel tubes with different diameters. Conventional B-mode and synthetic aperture images of the phantoms at different positions were obtained. The phantoms were automatically segmented from the ultrasound images using an ellipse fitting algorithm, the centroid of which is subsequently used as a fiducial for calibration. Calibration accuracy was evaluated for these procedures based on the leave-one-out target registration error. It was shown that larger diameter phantoms with lower

  5. Archimedean Spiral Antenna Calibration Procedures to Increase the Downrange Resolution of a SFCW Radar

    Directory of Open Access Journals (Sweden)

    Ioan Nicolaescu

    2008-01-01

    Full Text Available This paper deals with the calibration procedures of an Archimedean spiral antenna used for a stepped frequency continuous wave radar (SFCW, which works from 400 MHz to 4845 MHz. Two procedures are investigated, one based on an error-term flow graph for the frequency signal and the second based on a reference metallic plate located at a certain distance from the ground in order to identify the phase dispersion given by the antenna. In the second case, the received signal is passed in time domain by applying an ifft, the multiple reflections are removed and the phase variation due to the time propagation is subtracted. After phase correction, the time domain response as well as the side lobes level is decreased. The antenna system made up of two Archimedean spirals is employed by SFCW radar that operates with a frequency step of 35 MHz.

  6. Novel Real-time Calibration and Alignment Procedure for LHCb Run II

    CERN Multimedia

    Prouve, Claire

    2016-01-01

    In order to achieve optimal detector performance the LHCb experiment has introduced a novel real-time detector alignment and calibration strategy for Run II of the LHC. For the alignment tasks, data is collected and processed at the beginning of each fill while the calibrations are performed for each run. This real time alignment and calibration allows the same constants being used in both the online and offline reconstruction, thus improving the correlation between triggered and offline selected events. Additionally the newly computed alignment and calibration constants can be instantly used in the trigger, making it more efficient. The online alignment and calibration of the RICH detectors also enable the use of hadronic particle identification in the trigger. The computing time constraints are met through the use of a new dedicated framework using the multi-core farm infrastructure for the LHCb trigger. An overview of all alignment and calibration tasks is presented and their performance is shown.

  7. Improving transferability strategies for debris flow susceptibility assessment: Application to the Saponara and Itala catchments (Messina, Italy)

    KAUST Repository

    Cama, M.; Lombardo, Luigi; Conoscenti, C.; Rotigliano, E.

    2017-01-01

    provide an efficient solution to dealing with this limit. However, when applying a transferability procedure, a key point is the optimal selection of the predictors to be included for calibrating the model in the source area. In this paper, the issue

  8. Advancing Absolute Calibration for JWST and Other Applications

    Science.gov (United States)

    Rieke, George; Bohlin, Ralph; Boyajian, Tabetha; Carey, Sean; Casagrande, Luca; Deustua, Susana; Gordon, Karl; Kraemer, Kathleen; Marengo, Massimo; Schlawin, Everett; Su, Kate; Sloan, Greg; Volk, Kevin

    2017-10-01

    We propose to exploit the unique optical stability of the Spitzer telescope, along with that of IRAC, to (1) transfer the accurate absolute calibration obtained with MSX on very bright stars directly to two reference stars within the dynamic range of the JWST imagers (and of other modern instrumentation); (2) establish a second accurate absolute calibration based on the absolutely calibrated spectrum of the sun, transferred onto the astronomical system via alpha Cen A; and (3) provide accurate infrared measurements for the 11 (of 15) highest priority stars with no such data but with accurate interferometrically measured diameters, allowing us to optimize determinations of effective temperatures using the infrared flux method and thus to extend the accurate absolute calibration spectrally. This program is integral to plans for an accurate absolute calibration of JWST and will also provide a valuable Spitzer legacy.

  9. Electromagnetic Calorimeter Calibration with $\\pi^{0}$

    CERN Multimedia

    Puig Navarro, A

    2009-01-01

    Several methods can be used in order to achieve precise calibration of the LHCb Electromagnetic Calorimeter (ECAL) once reasonable cell equalization has been reached. At low transverse energy, the standard calibration procedure is an iterative method based on the fit of the $\\gamma\\gamma$ invariant mass distribution for each cell of the decay $\\pi^{0}\\to\\gamma\\gamma$ with resolved photons. A new technique for generating the combinatorial background of such decays directly from data has been developed. Knowledge of the background could allow an alternative calibration method based on a event by event fit of the same $\\gamma\\gamma$ invariant mass distribution where contributions from groups of cells are considered in a single fit. The background generation procedure and this possible new calibration method are presented in this poster, in addition to an overview of the LHCb Calorimetry system and ECAL calibration techniques.

  10. Mercury Continuous Emmission Monitor Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani

    2009-03-12

    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  11. Saved Leave Scheme (SLS) : Simplified procedure for the transfer of leave to saved leave accounts

    CERN Multimedia

    HR Division

    2001-01-01

    As part of the process of streamlining procedures, the HR and AS Divisions have jointly developed a system whereby annual and compensatory leave will henceforth be automatically transferred1) to saved leave accounts. Under the provisions of the voluntary saved leave scheme (SLS), a maximum total of 10 days'2) annual and compensatory leave (excluding saved leave accumulated in accordance with the provisions of Administrative Circular No. 22 B) can be transferred to the saved leave account at the end of the leave year (30 September). Previously, every person taking part in the scheme has been individually issued with a form for the purposes of requesting the transfer of leave to the leave account and the transfer has then had to be done manually by HR Division. To streamline the procedure, unused leave of all those taking part in the saved leave scheme at the closure of of the leave-year accounts will henceforth be transferred automatically to the saved leave account on that date. This simplification is in the ...

  12. Binary pseudo-random gratings and arrays for calibration of the modulation transfer function of surface profilometers: recent developments

    Energy Technology Data Exchange (ETDEWEB)

    Barber, Samuel K.; Soldate, Paul; Anderson, Erik H.; Cambie, Rossana; Marchesini, Stefano; McKinney, Wanye R.; Takacs, Peter Z.; Voronov, Dmitry L.; Yashchuk, Valeriy V.

    2009-07-07

    The major problem of measurement of a power spectral density (PSD) distribution of the surface heights with surface profilometers arises due to the unknown Modulation Transfer Function (MTF) of the instruments. The MTF tends to distort the PSD at higher spatial frequencies. It has been suggested [Proc. SPIE 7077-7, (2007), Opt. Eng. 47 (7), 073602-1-5 (2008)] that the instrumental MTF of a surface profiler can be precisely measured using standard test surfaces based on binary pseudo-random (BPR) patterns. In the cited work, a one dimensional (1D) realization of the suggested method based on use of BPR gratings has been demonstrated. Here, we present recent achievements made in fabricating and using two-dimensional (2D) BPR arrays that allow for a direct 2D calibration of the instrumental MTF. The 2D BPRAs were used as standard test surfaces for 2D MTF calibration of the MicromapTM-570 interferometric microscope with all available objectives. The effects of fabrication imperfections on the efficiency of calibration are also discussed.

  13. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 2: Data standardization for tank calibration

    International Nuclear Information System (INIS)

    2007-01-01

    Measurements of the volume and height of liquid in a process accountancy tank are often made in order to estimate or verify the tank's calibration or volume measurement equation. The calibration equation relates the response of the tank's measurement system to some independent measure of tank volume. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation (the inverse of the calibration equation), which relates tank volume to measurement system response. In this part of ISO 18213, it is assumed that the primary measurement-system response variable is liquid height and that the primary measure of liquid content is volume. This part of ISO 18213 presents procedures for standardizing a set of calibration data to a fixed set of reference conditions so as to minimize the effect of variations in ambient conditions that occur during the measurement process. The procedures presented herein apply generally to measurements of liquid height and volume obtained for the purpose of calibrating a tank (i.e. calibrating a tank's measurement system). When used in connection with other parts of ISO 18213, these procedures apply specifically to tanks equipped with bubbler probe systems for measuring liquid content. The standardization algorithms presented herein can be profitably applied when only estimates of ambient conditions, such as temperature, are available. However, the most reliable results are obtained when relevant ambient conditions are measured for each measurement of volume and liquid height in a set of calibration data. Information is provided on scope, physical principles, data required, calibration data, dimensional changes in the tank, multiple calibration runs and results on standardized calibration data. Four annexes inform about density of water, buoyancy corrections for mass determination, determination of tank heel volume and statistical method for aligning data from several calibration runs. A bibliography is

  14. Procedures for the definitive calibration of radiotherapy equipment

    International Nuclear Information System (INIS)

    1993-01-01

    Recommendations from the Institute of Physical Sciences in Medicine are given for the definition calibration of external beam radiotherapy treatment machines and radiation dose measuring equipment used in radiotherapy. (UK)

  15. Calibration of a detector by activation with a continuous neutron source used as a transfer standard for measuring pulsed neutron beams

    International Nuclear Information System (INIS)

    Moreno, Jose; Silva, Patricio; Birstein, Lipo; Soto, Leopoldo

    2002-01-01

    This paper presents a method for calibrating activation detectors. These detectors will be used as transfer standard in measuring neutron fluxes produced by pulsed plasma sources. A standard neutron source is used as a secondary standard. The activation detector is being shielded in order to substantially reduce detection of gamma emission coming from the source. The detector's calibration factor is obtained by considering also the standard neutron source as a free source of gamma radiation so that the measurements can be done without quickly withdrawing the neutron source as it is usually done. This will substantially simplify the traditionally established method (JM)

  16. Calibration guidelines for surface texture instruments - horizontal axis

    DEFF Research Database (Denmark)

    Andreasen, Jan Lasson; Shem, R. Krüger

    The present report is a documentation of the work carried out at DTU, on TASK 5.1: PROCEDURES FOR CALIBRATION IN X- AND Y- DIRECTION the project with contract SMT4-CT97-2176 with title: Calibration Standards for Surface Topography Measuring Systems down to Nanometric Scale. After a short introduc......The present report is a documentation of the work carried out at DTU, on TASK 5.1: PROCEDURES FOR CALIBRATION IN X- AND Y- DIRECTION the project with contract SMT4-CT97-2176 with title: Calibration Standards for Surface Topography Measuring Systems down to Nanometric Scale. After a short...

  17. Calibration and simulation of Heston model

    Directory of Open Access Journals (Sweden)

    Mrázek Milan

    2017-05-01

    Full Text Available We calibrate Heston stochastic volatility model to real market data using several optimization techniques. We compare both global and local optimizers for different weights showing remarkable differences even for data (DAX options from two consecutive days. We provide a novel calibration procedure that incorporates the usage of approximation formula and outperforms significantly other existing calibration methods.

  18. Optimized Spectrometers Characterization Procedure for Near Ground Support of ESA FLEX Observations: Part 1 Spectral Calibration and Characterisation

    Directory of Open Access Journals (Sweden)

    Laura Mihai

    2018-02-01

    Full Text Available The paper presents two procedures for the wavelength calibration, in the oxygen telluric absorption spectral bands (O2-A, λc = 687 nm and O2-B, λc = 760.6 nm, of field fixed-point spectrometers used for reflectance and Sun-induced fluorescence measurements. In the first case, Ne and Ar pen-type spectral lamps were employed, while the second approach is based on a double monochromator setup. The double monochromator system was characterized for the estimation of errors associated with different operating configurations. The proposed methods were applied to three Piccolo Doppio-type systems built around two QE Pros and one USB2 + H16355 Ocean Optics spectrometers. The wavelength calibration errors for all the calibrations performed on the three spectrometers are reported and potential methodological improvements discussed. The suggested calibration methods were validated, as the wavelength corrections obtained by both techniques for the QE Pro designed for fluorescence investigations were similar. However, it is recommended that a neon emission line source, as well as an argon or mercury-argon source be used to have a reference wavelength closer to the O2-B feature. The wavelength calibration can then be optimised as close to the O2-B and O2-A features as possible. The monochromator approach could also be used, but that instrument would need to be fully characterized prior to use, and although it may offer a more accurate calibration, as it could be tuned to emit light at the same wavelengths as the absorption features, it would be more time consuming as it is a scanning approach.

  19. Lidar to lidar calibration phase 1

    DEFF Research Database (Denmark)

    Yordanova, Ginka; Courtney, Michael

    This report presents a feasibility study of a lidar to lidar (L2L) calibration procedure. Phase one of the project was conducted at Høvsøre, Denmark. Two windcubes were placed next to the 116m met mast and different methods were applied to obtain the sensing height error of the lidars. The purpose...... is to find the most consistent method and use it in a potential lidar to lidar calibration procedure....

  20. Accurate calibration of waveform data measured by the Plasma Wave Experiment on board the ARASE satellite

    Science.gov (United States)

    Kitahara, M.; Katoh, Y.; Hikishima, M.; Kasahara, Y.; Matsuda, S.; Kojima, H.; Ozaki, M.; Yagitani, S.

    2017-12-01

    The Plasma Wave Experiment (PWE) is installed on board the ARASE satellite to measure the electric field in the frequency range from DC to 10 MHz, and the magnetic field in the frequency range from a few Hz to 100 kHz using two dipole wire-probe antennas (WPT) and three magnetic search coils (MSC), respectively. In particular, the Waveform Capture (WFC), one of the receivers of the PWE, can detect electromagnetic field waveform in the frequency range from a few Hz to 20 kHz. The Software-type Wave Particle Interaction Analyzer (S-WPIA) is installed on the ARASE satellite to measure the energy exchange between plasma waves and particles. Since S-WPIA uses the waveform data measured by WFC to calculate the relative phase angle between the wave magnetic field and velocity of energetic electrons, the high-accuracy is required to calibration of both amplitude and phase of the waveform data. Generally, the calibration procedure of the signal passed through a receiver consists of three steps; the transformation into spectra, the calibration by the transfer function of a receiver, and the inverse transformation of the calibrated spectra into the time domain. Practically, in order to reduce the side robe effect, a raw data is filtered by a window function in the time domain before applying Fourier transform. However, for the case that a first order differential coefficient of the phase transfer function of the system is not negligible, the phase of the window function convoluted into the calibrated spectra is shifted differently at each frequency, resulting in a discontinuity in the time domain of the calibrated waveform data. To eliminate the effect of the phase shift of a window function, we suggest several methods to calibrate a waveform data accurately and carry out simulations assuming simple sinusoidal waves as an input signal and using transfer functions of WPT, MSC, and WFC obtained in pre-flight tests. In consequence, we conclude that the following two methods can

  1. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 4: Accurate determination of liquid height in accountancy tanks equipped with dip tubes, slow bubbling rate

    International Nuclear Information System (INIS)

    2008-01-01

    reprocessing plants, the volume of liquid transferred into or out of a tank is determined by the levels of two siphons. The high level corresponds to the nominal volume, and the low level to the heel volume. If the transfer volume cannot be measured directly, then it is necessary to calibrate this volume because the difference between the actual volume and that used for inventory calculations will appear as a systematic error. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation, which relates tank volume to measurement system response. Steps for using the measurement equation to determine the volume of process liquid in the tank are presented in ISO 18213-1. The procedure presented in this part of ISO 18213 for determining the height of process liquid in a tank from a measurement of the pressure it induces in the tank's measurement system is also a key step in the procedure for determining process liquid volumes

  2. Nuclear fuel technology - Tank calibration and volume determination for nuclear materials accountancy - Part 5: Accurate determination of liquid height in accountancy tanks equipped with dip tubes, fast bubbling rate

    International Nuclear Information System (INIS)

    2008-01-01

    reprocessing plants, the volume of liquid transferred into or out of a tank is determined by the levels of two siphons. The high level corresponds to the nominal volume, and the low level to the heel volume. If the transfer volume cannot be measured directly, then it is necessary to calibrate this volume because the difference between the actual volume and that used for inventory calculations will appear as a systematic error. The ultimate purpose of the calibration exercise is to estimate the tank's volume measurement equation, which relates tank volume to measurement system response. Steps for using the measurement equation to determine the volume of process liquid in the tank are presented in ISO 18213-1. The procedure presented in this part of ISO 18213 for determining the height of process liquid in a tank from a measurement of the pressure it induces in the tank's measurement system is also a key step in the procedure for determining process liquid volumes

  3. Skew redundant MEMS IMU calibration using a Kalman filter

    International Nuclear Information System (INIS)

    Jafari, M; Sahebjameyan, M; Moshiri, B; Najafabadi, T A

    2015-01-01

    In this paper, a novel calibration procedure for skew redundant inertial measurement units (SRIMUs) based on micro-electro mechanical systems (MEMS) is proposed. A general model of the SRIMU measurements is derived which contains the effects of bias, scale factor error and misalignments. For more accuracy, the effect of lever arms of the accelerometers to the center of the table are modeled and compensated in the calibration procedure. Two separate Kalman filters (KFs) are proposed to perform the estimation of error parameters for gyroscopes and accelerometers. The predictive error minimization (PEM) stochastic modeling method is used to simultaneously model the effect of bias instability and random walk noise on the calibration Kalman filters to diminish the biased estimations. The proposed procedure is simulated numerically and has expected experimental results. The calibration maneuvers are applied using a two-axis angle turntable in a way that the persistency of excitation (PE) condition for parameter estimation is met. For this purpose, a trapezoidal calibration profile is utilized to excite different deterministic error parameters of the accelerometers and a pulse profile is used for the gyroscopes. Furthermore, to evaluate the performance of the proposed KF calibration method, a conventional least squares (LS) calibration procedure is derived for the SRIMUs and the simulation and experimental results compare the functionality of the two proposed methods with each other. (paper)

  4. Efficient mass calibration of magnetic sector mass spectrometers

    International Nuclear Information System (INIS)

    Roddick, J.C.

    1996-01-01

    Magnetic sector mass spectrometers used for automatic acquisition of precise isotopic data are usually controlled with Hall probes and software that uses polynomial equations to define and calibrate the mass-field relations required for mass focusing. This procedure requires a number of reference masses and careful tuning to define and maintain an accurate mass calibration. A simplified equation is presented and applied to several different magnetically controlled mass spectrometers. The equation accounts for nonlinearity in typical Hall probe controlled mass-field relations, reduces calibration to a linear fitting procedure, and is sufficiently accurate to permit calibration over a mass range of 2 to 200 amu with only two defining masses. Procedures developed can quickly correct for normal drift in calibrations and compensate for drift during isotopic analysis over a limited mass range such as a single element. The equation is: Field A·Mass 1/2 + B·(Mass) p where A, B, and p are constants. The power value p has a characteristic value for a Hall probe/controller and is insensitive to changing conditions, thus reducing calibration to a linear regression to determine optimum A and B. (author). 1 ref., 1 tab., 6 figs

  5. Is there inter-procedural transfer of skills in intraocular surgery?

    DEFF Research Database (Denmark)

    Thomsen, Ann Sofia Skou; Kiilgaard, Jens Folke; la Cour, Morten

    2017-01-01

    surgery on a virtual-reality simulator until passing a test with predefined validity evidence (cataract trainees) or to (2) no cataract surgery training (novices). Possible skill transfer was assessed using a test consisting of all 11 vitreoretinal modules on the EyeSi virtual-reality simulator. All...... participants repeated the test of vitreoretinal surgical skills until their performance curve plateaued. Three experienced vitreoretinal surgeons also performed the test to establish validity evidence. Analysis with independent samples t-tests was performed. RESULTS: The vitreoretinal test on the Eye......, p = 0.265), or maximum scores (785 ± 162 points versus 805 ± 73 points, p = 0.791). CONCLUSION: Pretraining in cataract surgery did not demonstrate any measurable effect on vitreoretinal procedural performance. The results of this study indicate that we should not anticipate extensive transfer...

  6. System for measuring the effect of fouling and corrosion on heat transfer under simulated OTEC conditions. [HTAU and LABTTF codes

    Energy Technology Data Exchange (ETDEWEB)

    Fetkovich, J.G.

    1976-12-01

    A complete system designed to measure, with high precision, changes in heat transfer rates due to fouling and corrosion of simulated heat exchanger tubes, at sea and under OTEC conditions is described. All aspects of the system are described in detail, including theory, mechanical design, electronics design, assembly procedures, test and calibration, operating procedures, laboratory results, field results, and data analysis programs.

  7. Fermentation process tracking through enhanced spectral calibration modeling.

    Science.gov (United States)

    Triadaphillou, Sophia; Martin, Elaine; Montague, Gary; Norden, Alison; Jeffkins, Paul; Stimpson, Sarah

    2007-06-15

    The FDA process analytical technology (PAT) initiative will materialize in a significant increase in the number of installations of spectroscopic instrumentation. However, to attain the greatest benefit from the data generated, there is a need for calibration procedures that extract the maximum information content. For example, in fermentation processes, the interpretation of the resulting spectra is challenging as a consequence of the large number of wavelengths recorded, the underlying correlation structure that is evident between the wavelengths and the impact of the measurement environment. Approaches to the development of calibration models have been based on the application of partial least squares (PLS) either to the full spectral signature or to a subset of wavelengths. This paper presents a new approach to calibration modeling that combines a wavelength selection procedure, spectral window selection (SWS), where windows of wavelengths are automatically selected which are subsequently used as the basis of the calibration model. However, due to the non-uniqueness of the windows selected when the algorithm is executed repeatedly, multiple models are constructed and these are then combined using stacking thereby increasing the robustness of the final calibration model. The methodology is applied to data generated during the monitoring of broth concentrations in an industrial fermentation process from on-line near-infrared (NIR) and mid-infrared (MIR) spectrometers. It is shown that the proposed calibration modeling procedure outperforms traditional calibration procedures, as well as enabling the identification of the critical regions of the spectra with regard to the fermentation process.

  8. Standardization of the calibration of brachytherapy sources at the IAEA dosimetry laboratory

    International Nuclear Information System (INIS)

    Shanta, A.; Andreo, P.

    1996-01-01

    A new service to SSDLs has been initiated at the IAEA Dosimetry Laboratory for providing calibrations of well-type ionisation chambers, used in brachytherapy applications, which are traceable to the International Measurement System. Considering that the most common radionuclide used in the developing countries is 137 Cs, two such sources of the type used for gynaecological intracavitary applications have been purchased by the Agency and calibrated at the National Institute of Standards and Technology (NIST), USA. These 137 Cs reference sources together with a well-type ionization chamber constitute the IAEA brachytherapy dosimetry standard. Based on the recommendations by a group of experts, a method has been developed for transferring calibrations to SSDLs which is described in this paper. The method is based on the acquisition by the SSDLs of sources and equipment similar to those at the IAEA. The well-type chamber is to be calibrated at the IAEA Dosimetry Laboratory, and this will be used at the SSDL to calibrate its own reference sources. These sources can in turn by used to calibrate well-type chambers from hospital users and to calibrate other type of sources by performing measurements in air. In order to standardize the procedures for the two methods and to provide guidance to the SSDLs, measurements have been carried out at the IAEA Dosimetry Laboratory. The reproducibility of the two type of measurements has been found to be better than 0.5%, and the uncertainty of calibrations estimated to be less than 1.5% (one standard deviation). (author). 8 refs, 8 figs, 2 tabs

  9. CryoSat-2 SIRAL Calibration: Strategy, Application and Results

    Science.gov (United States)

    Parrinello, T.; Fornari, M.; Bouzinac, C.; Scagliola, M.; Tagliani, N.

    2012-04-01

    The main payload of CryoSat-2 is a Ku band pulsewidth limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for azimuth processing. This allows to reach an along track resolution of about 250 meters which is an important improvement over traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed. In fact, not only the corrections for transfer function amplitude with respect to frequency, gain and instrument path delay have to be computed but it is also needed to provide corrections for transfer function phase with respect to frequency and AGC setting as well as the phase variation across bursts of pulses. As a consequence, SIRAL performs regularly four types of calibrations: (1) CAL1 in order to calibrate the internal path delay and peak power variation, (2) CAL2 in order to compensate the instrument transfer function, (3) CAL4 to calibrate the interferometer and (4) AutoCal, a specific sequence in order to calibrate the gain and phase difference for each AGC setting. Commissioning phase results (April-December 2010) revealed high stability of the instrument, which made possible to reduce the calibration frequency during Operations. Internal calibration data are processed on ground by the CryoSat-2 Instrument Processing Facility (IPF1) and then applied to the science data. In this poster we will describe as first the calibration strategy and then how the four different types of calibration are applied to science data. Moreover the calibration results over almost 2 years of mission will be presented, analyzing their temporal evolution in order to highlight the stability of the instrument over its life.

  10. Calibration of a Numerical Model for Heat Transfer and Fluid Flow in an Extruder

    DEFF Research Database (Denmark)

    Hofstätter, Thomas; Pedersen, David Bue; Nielsen, Jakob Skov

    2016-01-01

    This paper discusses experiments performed in order to validate simulations on a fused deposition modelling (FDM) extruder. The nozzle has been simulated in terms of heat transfer and fluid flow. In order to calibrate and validate these simulations, experiments were performed giving a significant...... look into the physical behaviour of the nozzle, heating and cooling systems. Experiments on the model were performed at different sub-mm diameters of the extruder. Physical parameters of the model – especially temperature dependent parameters – were set into analytical relationships in order to receive...... dynamical parameters. This research sets the foundation for further research within melted extrusion based additive manufacturing. The heating process of the extruder will be described and a note on the material feeding will be given....

  11. Implementation of ISO guide 25 in a medical dosimetry secondary standards calibration laboratory

    International Nuclear Information System (INIS)

    DeWerd, L.A.

    1995-01-01

    Currently, there is a great deal of discussion among industry and government agencies about ISO 9000 accreditation. U.S. manufacturers with ISO 9000 accreditation are regarded more favorably by European countries. The principles behind the ISO 9000 accreditation are based on the Total Quality Management (TQM) principles that are being implemented in many U.S. industries. This paper will deal only with the calibration issue. There is a difference in the areas covered by ISO 9000 and ISO Guide 25 documents. ISO 9000, in particular ISO 9001 - ISO 9003, cover the open-quotes calibrationclose quotes of inspection, measuring and test equipment. This equipment is basically used for open-quotes factory calibrationsclose quotes to determine that equipment is performing within manufacturer specifications. ISO Guide 25 is specifically for open-quotes calibration and testing laboratories,close quotes generally laboratories that have painstaking procedures to reduce uncertainties and establish high accuracy of the transfer of calibration. The experience of the University of Wisconsin Accredited Dosimetry Calibration Laboratory in conforming to ISO Guide 25 will be outlined. The entire laboratory staff must become familiar with the process and an individual with direct authority must become the one to maintain the quality of equipment and calibrations in the role of open-quotes quality-assurance manager.close quotes

  12. Test surfaces useful for calibration of surface profilometers

    Science.gov (United States)

    Yashchuk, Valeriy V; McKinney, Wayne R; Takacs, Peter Z

    2013-12-31

    The present invention provides for test surfaces and methods for calibration of surface profilometers, including interferometric and atomic force microscopes. Calibration is performed using a specially designed test surface, or the Binary Pseudo-random (BPR) grating (array). Utilizing the BPR grating (array) to measure the power spectral density (PSD) spectrum, the profilometer is calibrated by determining the instrumental modulation transfer.

  13. Simulating SiD Calorimetry: Software Calibration Procedures and Jet Energy Resolution

    International Nuclear Information System (INIS)

    Cassell, R.

    2009-01-01

    Simulated calorimeter performance in the SiD detector is examined. The software calibration procedures are described, as well as the perfect pattern recognition PFA reconstruction. Performance of the SiD calorimeters is summarized with jet energy resolutions from calorimetry only, perfect pattern recognition and the SiD PFA algorithm. Presented at LCWS08(1). Our objective is to simulate the calorimeter performance of the SiD detector, with and without a Particle Flow Algorithm (PFA). Full Geant4 simulations using SLIC(2) and the SiD simplified detector geometry (SiD02) are used. In this geometry, the calorimeters are represented as layered cylinders. The EM calorimeter is Si/W, with 20 layers of 2.5mm W and 10 layers of 5mm W, segmented in 3.5 x 3.5mm 2 cells. The HAD calorimeter is RPC/Fe, with 40 layers of 20mm Fe and a digital readout, segmented in 10 x 10mm 2 cells. The barrel detectors are layered in radius, while the endcap detectors are layered in z(along the beam axis)

  14. Using a heterodyne vibrometer in combination with pulse excitation for primary calibration of ultrasonic hydrophones in amplitude and phase

    Science.gov (United States)

    Weber, Martin; Wilkens, Volker

    2017-08-01

    A high-frequency vibrometer was used with ultrasonic pulse excitation in order to perform a primary hydrophone calibration. This approach enables the simultaneous characterization of the amplitude and phase transfer characteristic of ultrasonic hydrophones. The method allows a high frequency resolution in a considerably short time for the measurement. Furthermore, the uncertainty contributions of this approach were investigated and quantified. A membrane hydrophone was calibrated and the uncertainty budget for this measurement was determined. The calibration results are presented up to 70~\\text{MHz} . The measurement results show good agreement with the results obtained by sinusoidal burst excitation through the use of the vibrometer and by a homodyne laser interferometer, with RMS deviation of approximately 3% -4% in the frequency range from 1 to 60~\\text{MHz} . Further hydrophones were characterized up to 100~\\text{MHz} with this procedure to demonstrate the suitability for very high frequency calibration.

  15. Submicron beam X-ray diffraction of nanoheteroepitaxily grown GaN: Experimental challenges and calibration procedures

    Energy Technology Data Exchange (ETDEWEB)

    Bonanno, P.L., E-mail: PeterBonanno@gatech.ed [Georgia Institute of Technology/GTL, UMI 2958 Georgia Tech-CNRS, 57070 Metz (France); Gautier, S. [Laboratoire Materiaux Optiques, Photonique et Micro-Nano Systemes, UMR CNRS 7132, Universite de Metz et SUPELEC, 2 rue E. Belin, 57070 Metz (France); Sirenko, A.A. [Department of Physics, New Jersey Institute of Technology, Newark, NJ 07102 (United States); Kazimirov, A. [Cornell High Energy Synchrotron Source (CHESS), Cornell University, Ithaca, New York 14853 (United States); Cai, Z.-H. [Advanced Photon Source, 9700 S. Cass Avenue, Argonne, IL 60439 (United States); Goh, W.H. [Georgia Institute of Technology/GTL, UMI 2958 Georgia Tech-CNRS, 57070 Metz (France); Martin, J. [Laboratoire Materiaux Optiques, Photonique et Micro-Nano Systemes, UMR CNRS 7132, Universite de Metz et SUPELEC, 2 rue E. Belin, 57070 Metz (France); Martinez, A. [Laboratoire de Photonique et de Nanostructures, UPR CNRS 20, Route de Nozay, 91460 Marcoussis (France); Moudakir, T. [Laboratoire Materiaux Optiques, Photonique et Micro-Nano Systemes, UMR CNRS 7132, Universite de Metz et SUPELEC, 2 rue E. Belin, 57070 Metz (France); Maloufi, N. [Laboratoire d' Etude des Textures et Application aux Materiaux, UMR CNRS 7078, Ile du Saulcy, 57045 Metz Cedex 1 (France); Assouar, M.B. [Laboratoire de Physique des Milieux Ionises et Applications, Nancy University, CNRS, BP 239, F-54506 Vandoeuvre-les-Nancy Cedex (France); Ramdane, A.; Gratiet, L. Le [Laboratoire de Photonique et de Nanostructures, UPR CNRS 20, Route de Nozay, 91460 Marcoussis (France); Ougazzaden, A. [Georgia Institute of Technology/GTL, UMI 2958 Georgia Tech-CNRS, 57070 Metz (France)

    2010-02-15

    Highly relaxed GaN nanodots and submicron ridges have been selectively grown in the NSAG regime using MOVPE on lattice mismatched 6H-SiC and AlN substrates. 2D real space and 3D reciprocal space mapping was performed with a CCD detector using 10.4 keV synchrotron X-ray radiation at the 2-ID-D micro-diffraction beamline at Advanced Photon Source (APS). Calibration procedures have been developed to overcome the unique challenges of analyzing NSAG structures grown on highly mismatched substrates. We studied crystallographic planar bending on the submicron scale and found its correlation with strain relaxation in the NSAG ridges.

  16. Efficient mass calibration of magnetic sector mass spectrometers

    Energy Technology Data Exchange (ETDEWEB)

    Roddick, J C

    1997-12-31

    Magnetic sector mass spectrometers used for automatic acquisition of precise isotopic data are usually controlled with Hall probes and software that uses polynomial equations to define and calibrate the mass-field relations required for mass focusing. This procedure requires a number of reference masses and careful tuning to define and maintain an accurate mass calibration. A simplified equation is presented and applied to several different magnetically controlled mass spectrometers. The equation accounts for nonlinearity in typical Hall probe controlled mass-field relations, reduces calibration to a linear fitting procedure, and is sufficiently accurate to permit calibration over a mass range of 2 to 200 amu with only two defining masses. Procedures developed can quickly correct for normal drift in calibrations and compensate for drift during isotopic analysis over a limited mass range such as a single element. The equation is: Field A{center_dot}Mass{sup 1/2} + B{center_dot}(Mass){sup p} where A, B, and p are constants. The power value p has a characteristic value for a Hall probe/controller and is insensitive to changing conditions, thus reducing calibration to a linear regression to determine optimum A and B. (author). 1 ref., 1 tab., 6 figs.

  17. Instrumentation calibration

    International Nuclear Information System (INIS)

    Mack, D.A.

    1976-08-01

    Procedures for the calibration of different types of laboratory equipment are described. Provisions for maintaining the integrity of reference and working standards traceable back to a national standard are discussed. Methods of validation and certification methods are included. An appendix lists available publications and services of national standardizing agencies

  18. Calibration of well-type ionization chambers

    International Nuclear Information System (INIS)

    Alves, C.F.E.; Leite, S.P.; Pires, E.J.; Magalhaes, L.A.G.; David, M.G.; Almeida, C.E. de

    2015-01-01

    This paper presents the methodology developed by the Laboratorio de Ciencias Radiologicas and presently in use for determining of the calibration coefficient for well-type chambers used in the dosimetry of 192 Ir high dose rate sources. Uncertainty analysis involving the calibration procedure are discussed. (author)

  19. Quantum-chemistry based calibration of the alkali metal cation series (Li(+)-Cs(+)) for large-scale polarizable molecular mechanics/dynamics simulations.

    Science.gov (United States)

    Dudev, Todor; Devereux, Mike; Meuwly, Markus; Lim, Carmay; Piquemal, Jean-Philip; Gresh, Nohad

    2015-02-15

    The alkali metal cations in the series Li(+)-Cs(+) act as major partners in a diversity of biological processes and in bioinorganic chemistry. In this article, we present the results of their calibration in the context of the SIBFA polarizable molecular mechanics/dynamics procedure. It relies on quantum-chemistry (QC) energy-decomposition analyses of their monoligated complexes with representative O-, N-, S-, and Se- ligands, performed with the aug-cc-pVTZ(-f) basis set at the Hartree-Fock level. Close agreement with QC is obtained for each individual contribution, even though the calibration involves only a limited set of cation-specific parameters. This agreement is preserved in tests on polyligated complexes with four and six O- ligands, water and formamide, indicating the transferability of the procedure. Preliminary extensions to density functional theory calculations are reported. © 2014 Wiley Periodicals, Inc.

  20. The CVD graphene transfer procedure introduces metallic impurities which alter the graphene electrochemical properties.

    Science.gov (United States)

    Ambrosi, Adriano; Pumera, Martin

    2014-01-07

    High quality graphene films can be fabricated by chemical vapor deposition (CVD) using Ni and Cu as catalytic substrates. Such a synthesis procedure always requires a subsequent transfer process to be performed in order to eliminate the metallic substrate and transfer the graphene onto the desired surface. We show here that such a transfer process causes significant contamination of the graphene film with residual Fe and Ni metal impurities. Fe contamination derives from the use of Fe-based etching solutions to dissolve Ni (or Cu) substrates, while residual Ni (or Cu) is due to an incomplete metal substrate etching. The presence of these metallic impurities within the transferred graphene film affects tremendously its electrochemical behavior when adopted as an electrode material.

  1. Calibrations of a tritium extraction facility

    International Nuclear Information System (INIS)

    Bretscher, M.M.; Oliver, B.M.; Farrar, H. IV.

    1983-01-01

    A tritium extraction facility has been built for the purpose of measuring the absolute tritium concentration in neutron-irradiated lithium metal samples. Two independent calibration procedures have been used to determine what fraction, if any, of tritium is lost during the extraction process. The first procedure compares independently measured 4 He and 3 H concentrations from the 6 Li(n,α)T reaction. The second procedure compared measured 6 Li(n,α)T/ 197 Au (n,γ) 198 Au thermal neutron reaction rate ratios with those obtained from Monte Carlo calculations using well-known cross sections. Both calibration methods show that within experimental errors (approx. 1.5%) no tritium is lost during the extraction process

  2. SU-E-T-749: Thorough Calibration of MOSFET Dosimeters

    International Nuclear Information System (INIS)

    Plenkovich, D; Thomas, J

    2015-01-01

    Purpose: To improve the accuracy of the MOSFET calibration procedure by performing the measurement several times and calculating the average value of the calibration factor for various photon and electron energies. Methods: The output of three photon and six electron beams of Varian Trilogy linear accelerator SN 5878 was calibrated. Five reinforced standard sensitivity MOSFET dosimeters were placed in the calibration jig and connected to the Reader Module. As the backscatter material was used 7 cm of Virtual Water. The MOSFET dosimeters were covered with 1.5 cm thick bolus for the regular and SRS 6 MV beams, 3 cm bolus for 15 MV beam, 1.5 cm bolus for 6 MeV electron beam, and 2 cm bolus for the electron energies of 9, 12, 15, 18, and 22 MeV. The dosimeters were exposed to 100 MU, and the calibration factor was determined using the mobileMOSFET software. To improve the accuracy of calibration, this procedure was repeated ten times and the calibration factors were averaged. Results: As the number of calibrations was increasing the variability of calibration factors of different dosimeters was decreasing. After ten calibrations, the calibration factors for all five dosimeters were within 1% of one another for all energies, except 6 MV SRS photons and 6 MeV electrons, for which the variability was 2%. Conclusions: The described process results in calibration factors which are almost independent of modality or energy. Once calibrated, the dosimeters may be used for in-vivo dosimetry or for daily verification of the beam output. Measurement of the radiation dose under bolus and scatter to the eye are examples of frequent use of calibrated MOSFET dosimeters. The calibration factor determined for full build-up is used under these circumstances. To the best of our knowledge, such thorough procedure for calibrating MOSFET dosimeters has not been reported previously. Best Medical Canada provided MOSFET dosimeters for this project

  3. Physiotherapy ultrasound calibrations

    International Nuclear Information System (INIS)

    Gledhill, M.

    1996-01-01

    Calibration of physiotherapy ultrasound equipment has long been a problem. Numerous surveys around the world over the past 20 years have all found that only a low percentage of the units tested had an output within 30% of that indicatd. In New Zealand, a survey carried out by the NRL in 1985 found that only 24% had an output, at the maximum setting, within + or - 20% of that indicated. The present performance Standard for new equipment (NZS 3200.2.5:1992) requires that the measured output should not deviate from that indicated by more than + or - 30 %. This may be tightened to + or - 20% in the next few years. Any calibration is only as good as the calibration equipment. Some force balances can be tested with small weights to simulate the force exerted by an ultrasound beam, but with others this is not possible. For such balances, testing may only be feasible with a calibrated source which could be used like a transfer standard. (author). 4 refs., 3 figs

  4. Calibration and intercomparison methods of dose calibrators used in nuclear medicine facilities; Metodos de calibracao e de intercomparacao de calibradores de dose utilizados em servicos de medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Costa, Alessandro Martins da

    1999-07-01

    Dose calibrators are used in most of the nuclear medicine facilities to determine the amount of radioactivity administered to a patient in a particular investigation or therapeutic procedure. It is therefore of vital importance that the equipment used presents good performance and is regular;y calibrated at a authorized laboratory. This occurs of adequate quality assurance procedures are carried out. Such quality control tests should be performed daily, other biannually or yearly, testing, for example, its accuracy and precision, the reproducibility and response linearity. In this work a commercial dose calibrator was calibrated with solution of radionuclides used in nuclear medicine. Simple instrument tests, such as response linearity and the response variation of the source volume increase at a constant source activity concentration, were performed. This instrument can now be used as a working standard for calibration of other dose calibrators/ An intercomparison procedure was proposed as a method of quality control of dose calibrators used in nuclear medicine facilities. (author)

  5. Radiometric Cross-Calibration of GF-4 in Multispectral Bands

    Directory of Open Access Journals (Sweden)

    Aixia Yang

    2017-03-01

    Full Text Available The GaoFen-4 (GF-4, launched at the end of December 2015, is China’s first high-resolution geostationary optical satellite. A panchromatic and multispectral sensor (PMS is onboard the GF-4 satellite. Unfortunately, the GF-4 has no onboard calibration assembly, so on-orbit radiometric calibration is required. Like the charge-coupled device (CCD onboard HuanJing-1 (HJ or the wide field of view sensor (WFV onboard GaoFen-1 (GF-1, GF-4 also has a wide field of view, which provides challenges for cross-calibration with narrow field of view sensors, like the Landsat series. A new technique has been developed and used to calibrate HJ-1/CCD and GF-1/WFV, which is verified viable. The technique has three key steps: (1 calculate the surface using the bi-directional reflectance distribution function (BRDF characterization of a site, taking advantage of its uniform surface material and natural topographic variation using Landsat Enhanced Thematic Mapper Plus (ETM+/Operational Land Imager (OLI imagery and digital elevation model (DEM products; (2 calculate the radiance at the top-of-the atmosphere (TOA with the simulated surface reflectance using the atmosphere radiant transfer model; and (3 fit the calibration coefficients with the TOA radiance and corresponding Digital Number (DN values of the image. This study attempts to demonstrate the technique is also feasible to calibrate GF-4 multispectral bands. After fitting the calibration coefficients using the technique, extensive validation is conducted by cross-validation using the image pairs of GF-4/PMS and Landsat-8/OLI with similar transit times and close view zenith. The validation result indicates a higher accuracy and frequency than that given by the China Centre for Resources Satellite Data and Application (CRESDA using vicarious calibration. The study shows that the new technique is also quite feasible for GF-4 multispectral bands as a routine long-term procedure.

  6. 40 CFR 1065.315 - Pressure, temperature, and dewpoint calibration.

    Science.gov (United States)

    2010-07-01

    ...) AIR POLLUTION CONTROLS ENGINE-TESTING PROCEDURES Calibrations and Verifications Measurement of Engine... temperature-equilibrated and temperature-monitored calibration salt solutions in containers that seal...

  7. UV scale calibration transfer from an improved pyroelectric detector standard to field UV-A meters and 365 nm excitation sources

    Science.gov (United States)

    Eppeldauer, G. P.; Podobedov, V. B.; Cooksey, C. C.

    2017-05-01

    Calibration of the emitted radiation from UV sources peaking at 365 nm, is necessary to perform the ASTM required 1 mW/cm2 minimum irradiance in certain military material (ships, airplanes etc) tests. These UV "black lights" are applied for crack-recognition using fluorescent liquid penetrant inspection. At present, these nondestructive tests are performed using Hg-lamps. Lack of a proper standard and the different spectral responsivities of the available UV meters cause significant measurement errors even if the same UV-365 source is measured. A pyroelectric radiometer standard with spectrally flat (constant) response in the UV-VIS range has been developed to solve the problem. The response curve of this standard determined from spectral reflectance measurement, is converted into spectral irradiance responsivity with UV sources (with different peaks and distributions) without using any source standard. Using this broadband calibration method, yearly spectral calibrations for the reference UV (LED) sources and irradiance meters is not needed. Field UV sources and meters can be calibrated against the pyroelectric radiometer standard for broadband (integrated) irradiance and integrated responsivity. Using the broadband measurement procedure, the UV measurements give uniform results with significantly decreased uncertainties.

  8. Increased Automation in Stereo Camera Calibration Techniques

    Directory of Open Access Journals (Sweden)

    Brandi House

    2006-08-01

    Full Text Available Robotic vision has become a very popular field in recent years due to the numerous promising applications it may enhance. However, errors within the cameras and in their perception of their environment can cause applications in robotics to fail. To help correct these internal and external imperfections, stereo camera calibrations are performed. There are currently many accurate methods of camera calibration available; however, most or all of them are time consuming and labor intensive. This research seeks to automate the most labor intensive aspects of a popular calibration technique developed by Jean-Yves Bouguet. His process requires manual selection of the extreme corners of a checkerboard pattern. The modified process uses embedded LEDs in the checkerboard pattern to act as active fiducials. Images are captured of the checkerboard with the LEDs on and off in rapid succession. The difference of the two images automatically highlights the location of the four extreme corners, and these corner locations take the place of the manual selections. With this modification to the calibration routine, upwards of eighty mouse clicks are eliminated per stereo calibration. Preliminary test results indicate that accuracy is not substantially affected by the modified procedure. Improved automation to camera calibration procedures may finally penetrate the barriers to the use of calibration in practice.

  9. User-calibration of Mettler AT200 analytical balance

    International Nuclear Information System (INIS)

    Estill, J.

    1996-01-01

    The purpose of this technical implementing procedure (TIP) is to describe the calibration of the Mettler AT200 analytical balance or similar type balance (henceforth called the balance). This balance is used for activities of the Scientific Investigation Plan (SIP) ''Metal Barrier Selection and Testing'' (SIP-CM-01, WBS nr. 1.2.2.5.1). In particular, it will be used for Activity E-20-50, ''Long-Term Corrosion Studies.'' The balance will be used for weighing test specimens and reagent chemicals. However, it is not limited to these uses. The calibration procedures consist of activating the internal (self) calibration of the apparatus, and weighing and recording of traceable standards. The balance is equipped with self (internal) calibration and linearization capabilities. It has an internal (built in) set of weights which are used for self calibration. The standard weights are traceable to National Institute of Standards and Technology (NIST)

  10. Application of Quasi-Heat-Pulse Solutions for Luikov’s Equations of Heat and Moisture Transfer for Calibrating and Utilizing Thermal Properties Apparatus

    Science.gov (United States)

    Mark A. Dietenberger; Charles R. Boardman

    2014-01-01

    Several years ago the Laplace transform solutions of Luikov’s differential equations were presented for one-dimensional heat and moisture transfer in porous hydroscopic orthotropic materials for the boundary condition of a gradual heat pulse applied to both surfaces of a flat slab. This paper presents calibration methods and data for the K-tester 637 (Lasercomp),...

  11. Development of an in situ calibration technique for combustible gas detectors

    Science.gov (United States)

    Shumar, J. W.; Wynveen, R. A.; Lance, N., Jr.; Lantz, J. B.

    1977-01-01

    This paper describes the development of an in situ calibration procedure for combustible gas detectors (CGD). The CGD will be a necessary device for future space vehicles as many subsystems in the Environmental Control/Life Support System utilize or produce hydrogen (H2) gas. Existing calibration techniques are time-consuming and require support equipment such as an environmental chamber and calibration gas supply. The in situ calibration procedure involves utilization of a water vapor electrolysis cell for the automatic in situ generation of a H2/air calibration mixture within the flame arrestor of the CGD. The development effort concluded with the successful demonstration of in situ span calibrations of a CGD.

  12. Method of calibration for portable detectors of ionizing radiation

    International Nuclear Information System (INIS)

    Leal, Carlos; Carrizales, Lila; Guacaran, Douglas; Moreno, Ailed; Duran, Jose

    2008-01-01

    Full text: The LSCD is beginning a process of accreditation by the national authority of Venezuela (CENCAMER), so we felt the need to modify existing protocols calibration, suggested by the IAEA (Technical Reports Series No 133) to be adjusted in accordance with international standards of ISO 17025 EA-4/02 and Mexican standard NOM-021-NUCL-2002, adapting the calibration procedure of portable radiation detectors to those standards. The procedure used in the LSCD, conducts an intercomparison between measures between the standard reference and to calibrate the instrument, thus verifying the correct operation of the equipment and its usefulness to the individual or environmental radiation monitoring, not to exceed the limits primary for the public and occupationally Staff Exposed (POE), and propose a simple way to perform this calibration procedure and comply with the standard international and national standards, ensuring the proper functioning of the instruments calibrated. One of the calculations that has proven to be a useful tool in the verification of the conditions of operating the same, it is uncertainty, this procedure allows to characterize the response of individual detector, depending on the constraints specified by the manufacturer, to optimize the calibration is necessary to consider the multiple factors that can affect the measure and evaluate the response of the same, so it is vitally important estimating the typical uncertainty associated with the calibration technique employed, but this calculation is not trivial, depends on the type of detector found usually between 5% to 20%, both in uncertainty as the linearity, which are the parameters that are set to consider an instrument in good or bad condition. Typical are calibrated are: monitors air Geiger-Muller, Ionization cameras, detector of Neutron's, Proportional Counter, Personal dosimeters. (author)

  13. Chaos, Consternation and CALIPSO Calibration: New Strategies for Calibrating the CALIOP 1064 nm Channel

    Science.gov (United States)

    Vaughan, Mark; Garnier, Anne; Liu, Zhaoyan; Josset, Damien; Hu, Yongxiang; Lee, Kam-Pui; Hunt, William; Vernier, Jean-Paul; Rodier, Sharon; Pelon, Jaques; hide

    2012-01-01

    The very low signal-to-noise ratios of the 1064 nm CALIOP molecular backscatter signal make it effectively impossible to employ the "clear air" normalization technique typically used to calibrate elastic back-scatter lidars. The CALIPSO mission has thus chosen to cross-calibrate their 1064 nm measurements with respect to the 532 nm data using the two-wavelength backscatter from cirrus clouds. In this paper we discuss several known issues in the version 3 CALIOP 1064 nm calibration procedure, and describe the strategies that will be employed in the version 4 data release to surmount these problems.

  14. Multi-Sensor Calibration of Low-Cost Magnetic, Angular Rate and Gravity Systems

    Directory of Open Access Journals (Sweden)

    Markus Lüken

    2015-10-01

    Full Text Available We present a new calibration procedure for low-cost nine degrees-of-freedom (9DOF magnetic, angular rate and gravity (MARG sensor systems, which relies on a calibration cube, a reference table and a body sensor network (BSN. The 9DOF MARG sensor is part of our recently-developed “Integrated Posture and Activity Network by Medit Aachen” (IPANEMA BSN. The advantage of this new approach is the use of the calibration cube, which allows for easy integration of two sensor nodes of the IPANEMA BSN. One 9DOF MARG sensor node is thereby used for calibration; the second 9DOF MARG sensor node is used for reference measurements. A novel algorithm uses these measurements to further improve the performance of the calibration procedure by processing arbitrarily-executed motions. In addition, the calibration routine can be used in an alignment procedure to minimize errors in the orientation between the 9DOF MARG sensor system and a motion capture inertial reference system. A two-stage experimental study is conducted to underline the performance of our calibration procedure. In both stages of the proposed calibration procedure, the BSN data, as well as reference tracking data are recorded. In the first stage, the mean values of all sensor outputs are determined as the absolute measurement offset to minimize integration errors in the derived movement model of the corresponding body segment. The second stage deals with the dynamic characteristics of the measurement system where the dynamic deviation of the sensor output compared to a reference system is Sensors 2015, 15 25920 corrected. In practical validation experiments, this procedure showed promising results with a maximum RMS error of 3.89°.

  15. Multi-sensor calibration of low-cost magnetic, angular rate and gravity systems.

    Science.gov (United States)

    Lüken, Markus; Misgeld, Berno J E; Rüschen, Daniel; Leonhardt, Steffen

    2015-10-13

    We present a new calibration procedure for low-cost nine degrees-of-freedom (9DOF) magnetic, angular rate and gravity (MARG) sensor systems, which relies on a calibration cube, a reference table and a body sensor network (BSN). The 9DOF MARG sensor is part of our recently-developed "Integrated Posture and Activity Network by Medit Aachen" (IPANEMA) BSN. The advantage of this new approach is the use of the calibration cube, which allows for easy integration of two sensor nodes of the IPANEMA BSN. One 9DOF MARG sensor node is thereby used for calibration; the second 9DOF MARG sensor node is used for reference measurements. A novel algorithm uses these measurements to further improve the performance of the calibration procedure by processing arbitrarily-executed motions. In addition, the calibration routine can be used in an alignment procedure to minimize errors in the orientation between the 9DOF MARG sensor system and a motion capture inertial reference system. A two-stage experimental study is conducted to underline the performance of our calibration procedure. In both stages of the proposed calibration procedure, the BSN data, as well as reference tracking data are recorded. In the first stage, the mean values of all sensor outputs are determined as the absolute measurement offset to minimize integration errors in the derived movement model of the corresponding body segment. The second stage deals with the dynamic characteristics of the measurement system where the dynamic deviation of the sensor output compared to a reference system is Sensors 2015, 15 25920 corrected. In practical validation experiments, this procedure showed promising results with a maximum RMS error of 3.89°.

  16. Vibration transducer calibration techniques

    Science.gov (United States)

    Brinkley, D. J.

    1980-09-01

    Techniques for the calibration of vibration transducers used in the Aeronautical Quality Assurance Directorate of the British Ministry of Defence are presented. Following a review of the types of measurements necessary in the calibration of vibration transducers, the performance requirements of vibration transducers, which can be used to measure acceleration, velocity or vibration amplitude, are discussed, with particular attention given to the piezoelectric accelerometer. Techniques for the accurate measurement of sinusoidal vibration amplitude in reference-grade transducers are then considered, including the use of a position sensitive photocell and the use of a Michelson laser interferometer. Means of comparing the output of working-grade accelerometers with that of previously calibrated reference-grade devices are then outlined, with attention given to a method employing a capacitance bridge technique and a method to be used at temperatures between -50 and 200 C. Automatic calibration procedures developed to speed up the calibration process are outlined, and future possible extensions of system software are indicated.

  17. Calibration of areal surface topography measuring instruments

    Science.gov (United States)

    Seewig, J.; Eifler, M.

    2017-06-01

    The ISO standards which are related to the calibration of areal surface topography measuring instruments are the ISO 25178-6xx series which defines the relevant metrological characteristics for the calibration of different measuring principles and the ISO 25178-7xx series which defines the actual calibration procedures. As the field of areal measurement is however not yet fully standardized, there are still open questions to be addressed which are subject to current research. Based on this, selected research results of the authors in this area are presented. This includes the design and fabrication of areal material measures. For this topic, two examples are presented with the direct laser writing of a stepless material measure for the calibration of the height axis which is based on the Abbott- Curve and the manufacturing of a Siemens star for the determination of the lateral resolution limit. Based on these results, as well a new definition for the resolution criterion, the small scale fidelity, which is still under discussion, is presented. Additionally, a software solution for automated calibration procedures is outlined.

  18. A Room-Temperature Pre-calibration Procedure for Gradiometer Sifting

    International Nuclear Information System (INIS)

    Zhang Shu-Lin; Liu Yang-Bo; Wang Yong-Liang; Kong Xiang-Yan; Xie Xiao-Ming; Liu Ming

    2011-01-01

    In order to detect extremely weak magnetic signals, superconducting quantum interference device (SQUID) gradiometers are widely used to suppress environmental noise. A hardware SQUID gradiometer consists of a niobium gradio-antenna and an SQUID, which are coupled via an input coil. Here gradiometer imbalance may greatly reduce its noise suppression performance. The gradiometer balance depends on the geometrical forms of the antenna wound by niobium wire. We describe a simple method based on Faraday's law for the pre-calibration of the gradiometer balance at room temperature, before the gradiometer is set up. The pre-calibrating results are compared with the measured balance of an SQUID gradiometer system. This method may be used for sifting hardware gradiometers for multi-channel systems. (cross-disciplinary physics and related areas of science and technology)

  19. Cuban-Brazilian comparison of the calibration procedures for surface contamination monitors

    International Nuclear Information System (INIS)

    Salas, Gonzalo W.; Garcia, Jose A.T.; Ramos, Manoel M.O.

    2007-01-01

    The calibration laboratory of the CPHR, Centro de Proteccion e Higiene de las Radiaciones, is the only laboratory that performs calibrations of radiation protection instruments in Cuba. Recently they started a cooperation project with IAEA, code number CUB3002-01 with the task title 'Assuring and Demonstrating the safety of radioactive Waste Management'. Within the frame of this project they have a compromise to implement the calibration service for surface contamination monitors, for which they received instruments and sources. In 2006 the Laboratorio Nacional de Metrologia das Radiacoes Ionizantes (LNMRI) from the Instituto de Radioprotecao e Dosimetria, IRD/CNEN, was invited to perform a mission to CPHR in order to evaluate the level of implementation of the service trough the insertion of the service in the Quality Assurance System of their laboratory, and perform a comparison exercise to test the calibration method implemented. The results were evaluated by the methodology recommended in the ISO/IEC 43-1 standard. These results show that all values from both laboratories are comparable for most radionuclides except for Cl-36. The preliminary investigation about this situation is presented in the paper. (author)

  20. 40 CFR 92.122 - Smoke meter calibration.

    Science.gov (United States)

    2010-07-01

    ... 40 Protection of Environment 20 2010-07-01 2010-07-01 false Smoke meter calibration. 92.122 Section 92.122 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY (CONTINUED) AIR PROGRAMS... meter calibration. The smokemeter shall be checked according to the following procedure prior to each...

  1. Effects of line fiducial parameters and beamforming on ultrasound calibration.

    Science.gov (United States)

    Ameri, Golafsoun; Baxter, John S H; McLeod, A Jonathan; Peters, Terry M; Chen, Elvis C S

    2017-01-01

    Ultrasound (US)-guided interventions are often enhanced via integration with an augmented reality environment, a necessary component of which is US calibration. Calibration requires the segmentation of fiducials, i.e., a phantom, in US images. Fiducial localization error (FLE) can decrease US calibration accuracy, which fundamentally affects the total accuracy of the interventional guidance system. Here, we investigate the effects of US image reconstruction techniques as well as phantom material and geometry on US calibration. It was shown that the FLE was reduced by 29% with synthetic transmit aperture imaging compared with conventional B-mode imaging in a Z-bar calibration, resulting in a 10% reduction of calibration error. In addition, an evaluation of a variety of calibration phantoms with different geometrical and material properties was performed. The phantoms included braided wire, plastic straws, and polyvinyl alcohol cryogel tubes with different diameters. It was shown that these properties have a significant effect on calibration error, which is a variable based on US beamforming techniques. These results would have important implications for calibration procedures and their feasibility in the context of image-guided procedures.

  2. Calibrating a tensor magnetic gradiometer using spin data

    Science.gov (United States)

    Bracken, Robert E.; Smith, David V.; Brown, Philip J.

    2005-01-01

    Scalar magnetic data are often acquired to discern characteristics of geologic source materials and buried objects. It is evident that a great deal can be done with scalar data, but there are significant advantages to direct measurement of the magnetic gradient tensor in applications with nearby sources, such as unexploded ordnance (UXO). To explore these advantages, we adapted a prototype tensor magnetic gradiometer system (TMGS) and successfully implemented a data-reduction procedure. One of several critical reduction issues is the precise determination of a large group of calibration coefficients for the sensors and sensor array. To resolve these coefficients, we devised a spin calibration method, after similar methods of calibrating space-based magnetometers (Snare, 2001). The spin calibration procedure consists of three parts: (1) collecting data by slowly revolving the sensor array in the Earth?s magnetic field, (2) deriving a comprehensive set of coefficients from the spin data, and (3) applying the coefficients to the survey data. To show that the TMGS functions as a tensor gradiometer, we conducted an experimental survey that verified that the reduction procedure was effective (Bracken and Brown, in press). Therefore, because it was an integral part of the reduction, it can be concluded that the spin calibration was correctly formulated with acceptably small errors.

  3. High-dose secondary calibration laboratory accreditation program

    Energy Technology Data Exchange (ETDEWEB)

    Humphreys, J.C. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program.

  4. High-dose secondary calibration laboratory accreditation program

    International Nuclear Information System (INIS)

    Humphreys, J.C.

    1993-01-01

    There is a need for high-dose secondary calibration laboratories to serve the multi-billion dollar radiation processing industry. This need is driven by the desires of industry for less costly calibrations and faster calibration-cycle response time. Services needed include calibration irradiations of routine processing dosimeters and the supply of reference standard transfer dosimeters for irradiation in the production processing facility. In order to provide measurement quality assurance and to demonstrate consistency with national standards, the high-dose secondary laboratories would be accredited by means of an expansion of an existing National Voluntary Laboratory Accreditation Program. A laboratory performance criteria document is under development to implement the new program

  5. Instructional strategy effects on the retention and transfer of procedures of different difficulty level

    NARCIS (Netherlands)

    Jelsma, Otto; Pieters, Julius Marie

    1989-01-01

    In the present study, the effects of two instructional strategies on the retention and transfer of procedures of different difficulty level were investigated. Difficulty level was manipulated by providing a different number of cues during training. The instructional strategies differed with respect

  6. Calibration of hydrometers

    Science.gov (United States)

    Lorefice, Salvatore; Malengo, Andrea

    2006-10-01

    After a brief description of the different methods employed in periodic calibration of hydrometers used in most cases to measure the density of liquids in the range between 500 kg m-3 and 2000 kg m-3, particular emphasis is given to the multipoint procedure based on hydrostatic weighing, known as well as Cuckow's method. The features of the calibration apparatus and the procedure used at the INRiM (formerly IMGC-CNR) density laboratory have been considered to assess all relevant contributions involved in the calibration of different kinds of hydrometers. The uncertainty is strongly dependent on the kind of hydrometer; in particular, the results highlight the importance of the density of the reference buoyant liquid, the temperature of calibration and the skill of operator in the reading of the scale in the whole assessment of the uncertainty. It is also interesting to realize that for high-resolution hydrometers (division of 0.1 kg m-3), the uncertainty contribution of the density of the reference liquid is the main source of the total uncertainty, but its importance falls under about 50% for hydrometers with a division of 0.5 kg m-3 and becomes somewhat negligible for hydrometers with a division of 1 kg m-3, for which the reading uncertainty is the predominant part of the total uncertainty. At present the best INRiM result is obtained with commercially available hydrometers having a scale division of 0.1 kg m-3, for which the relative uncertainty is about 12 × 10-6.

  7. Development of a correction factor for Xe-133 vials for use with a dose calibrator

    International Nuclear Information System (INIS)

    Gels, G.L.; Piltingsrud, H.V.

    1982-01-01

    Manufacturers of dose calibrators who give calibration settings for various radionuclies sometimes do not specify the type of radionuclide container the calibration is for. The container, moreover, may not be of the same type as those a user might purchase. When these factors are not considered, the activity administered to the patient may be significantly different from that intended. An experiment is described in which calibration factors are determined for measurement of Xe-133 activity in vials in a dose calibrator. This was accomplished by transferring the Xe-133 from the commercial vials to standard NBS calibration ampules. Based on ten such transfers, the resulting correction factor for the dose calibrator was 1.22

  8. Calibration method of radiation monitoring system at TQNPC

    International Nuclear Information System (INIS)

    Liu Zhengshan; Zhang Qingli; Liu Jinjin; Miao Yuxing; Geng Lixin; Zhuang Yun; Dong Jianfeng; He Change

    2009-04-01

    The calibration methods and calibration device for standard monitor of radioactive particulate, iodine, noble gas and so on are not yet set up at home. On consideration of the present situation of the radiation monitoring system at the Third Qinshan Nuclear Power Co. Ltd., we have studied the calibration method of these radiation monitoring instruments used for measuring the waste liquid, particulate, iodine and noble gas produced during the operation of nuclear reactor. Through the check against these instruments during the No. 202 and No. 103 overhaul, we got initially the method of the calibration and obtained the transfer coefficient of calibration when secondary solid sources are used for calibration. Through the testing and calibration, the credibility of the radiation monitoring system is enhanced. And at the same time, the problems existing in the calibration are discussed. (authors)

  9. Radioactivity measurement of 18F in 16 ml vials for calibration of radionuclide calibrators

    International Nuclear Information System (INIS)

    Wurdiyanto, Gatot; Marsoem, Pujadi; Candra, Hermawan; Wijono, Paidi

    2012-01-01

    Fluorine-18 is obtained through the reaction 18 O(p, n) 18 F using a cyclotron that is situated in a hospital in Jakarta. Standardization of the 18 F solution is performed by gamma spectrometry using calibration sources of 152 Eu, 60 Co and 137 Cs that have traceability to the International System of units (SI). The activities in the 16 ml vials that were used for calibrating the radionuclide calibrators were between 1 and 2 GBq, with expanded uncertainties of 3.8%. The expanded uncertainty, at a coverage factor of k=2, on the derived calibration factor for the radionuclide calibrator was 6.6%. - Highlights: ► PTKMR–BATAN as a NMI of Indonesia is required to have procedures to calibrate the radionuclide calibrators. ► Standardizations were carried out on a solution of [ 18 F]FDG using gamma spectrometry. ► The volume of 18 F solutions used was 16 ml because this is the volume often used in hospitals. ► The Secondary Standard ionization chamber is a CRC-7BT Capintec radionuclide calibrator. ► A dial setting for 16 ml of [ 18 F]FDG solution in a vial is 443 for the Capintec dose calibrator.

  10. Reliability-Based Code Calibration

    DEFF Research Database (Denmark)

    Faber, M.H.; Sørensen, John Dalsgaard

    2003-01-01

    The present paper addresses fundamental concepts of reliability based code calibration. First basic principles of structural reliability theory are introduced and it is shown how the results of FORM based reliability analysis may be related to partial safety factors and characteristic values....... Thereafter the code calibration problem is presented in its principal decision theoretical form and it is discussed how acceptable levels of failure probability (or target reliabilities) may be established. Furthermore suggested values for acceptable annual failure probabilities are given for ultimate...... and serviceability limit states. Finally the paper describes the Joint Committee on Structural Safety (JCSS) recommended procedure - CodeCal - for the practical implementation of reliability based code calibration of LRFD based design codes....

  11. Calibration Methods for Reliability-Based Design Codes

    DEFF Research Database (Denmark)

    Gayton, N.; Mohamed, A.; Sørensen, John Dalsgaard

    2004-01-01

    The calibration methods are applied to define the optimal code format according to some target safety levels. The calibration procedure can be seen as a specific optimization process where the control variables are the partial factors of the code. Different methods are available in the literature...

  12. The METAS absorbed dose to water calibration service for high energy photon and electron beam radiotherapy

    International Nuclear Information System (INIS)

    Stucki, G.; Muench, W.; Quintel, H.

    2002-01-01

    mass. This experiment is similar to the one described by Feist, but extended to an energy range from 5.3 MeV to 22.4 MeV, allowing to determine the energy dependence of the response of the Fricke dosimeter. The absorbed dose to Fricke solution is determined using the particle energy, the total beam charge and the mass of the solution. The absorbed dose to Fricke solution is converted to an absorbed dose to water applying a general conversion factor taken from Ma et al. The thus calibrated Fricke solution is then used to calibrate several METAS plane-parallel transfer ionisation chambers of type NACP-02 in the mentioned energy range. The user dosimeters are finally compared to the METAS transfer standards following the procedures described in IAEA Technical Reports Series No. 398. It is anticipated that the overall uncertainty in the calibration factor of a user dosimeter will be around 2% (coverage factor k=2). (author)

  13. Validating and comparing GNSS antenna calibrations

    Science.gov (United States)

    Kallio, Ulla; Koivula, Hannu; Lahtinen, Sonja; Nikkonen, Ville; Poutanen, Markku

    2018-03-01

    GNSS antennas have no fixed electrical reference point. The variation of the phase centre is modelled and tabulated in antenna calibration tables, which include the offset vector (PCO) and phase centre variation (PCV) for each frequency according to the elevations and azimuths of the incoming signal. Used together, PCV and PCO reduce the phase observations to the antenna reference point. The remaining biases, called the residual offsets, can be revealed by circulating and rotating the antennas on pillars. The residual offsets are estimated as additional parameters when combining the daily GNSS network solutions with full covariance matrix. We present a procedure for validating the antenna calibration tables. The dedicated test field, called Revolver, was constructed at Metsähovi. We used the procedure to validate the calibration tables of 17 antennas. Tables from the IGS and three different calibration institutions were used. The tests show that we were able to separate the residual offsets at the millimetre level. We also investigated the influence of the calibration tables from the different institutions on site coordinates by performing kinematic double-difference baseline processing of the data from one site with different antenna tables. We found small but significant differences between the tables.

  14. STAR barrel electromagnetic calorimeter absolute calibration using 'minimum ionizing particles' from collisions at RHIC

    International Nuclear Information System (INIS)

    Cormier, T.M.; Pavlinov, A.I.; Rykov, M.V.; Rykov, V.L.; Shestermanov, K.E.

    2002-01-01

    The procedure for the STAR Barrel Electromagnetic Calorimeter (BEMC) absolute calibrations, using penetrating charged particle hits (MIP-hits) from physics events at RHIC, is presented. Its systematic and statistical errors are evaluated. It is shown that, using this technique, the equalization and transfer of the absolute scale from the test beam can be done to a percent level accuracy in a reasonable amount of time for the entire STAR BEMC. MIP-hits would also be an effective tool for continuously monitoring the variations of the BEMC tower's gains, virtually without interference to STAR's main physics program. The method does not rely on simulations for anything other than geometric and some other small corrections, and also for estimations of the systematic errors. It directly transfers measured test beam responses to operations at RHIC

  15. Validation of calibration procedures for freeform parts on CMMs

    DEFF Research Database (Denmark)

    Savio, Enrico; De Chiffre, Leonardo

    2003-01-01

    The paper describes the validation of a new method for establishment of traceability of freeform measurements on coordinate measuring machines currently being considered for development as a new ISO standard. The method deals with calibration by: i) repeated measurements of a given uncalibrated...

  16. Evaluation of uncertainties in the calibration of radiation survey meter

    International Nuclear Information System (INIS)

    Potiens, M.P.A.; Santos, G.P.

    2006-01-01

    In order to meet the requirements of ISO 17025, the quantification of the expanded uncertainties of experimental data in the calibration of survey meters must be carried out using well defined concepts, like those expressed in the 'ISO-Guide to the Expression of Uncertainty in Measurement'. The calibration procedure of gamma ray survey meters involves two values that have to get their uncertainties clearly known: measurements of the instrument under calibration and the conventional true values of a quantity. Considering the continuous improvement of the calibration methods and set-ups, it is necessary to evaluate periodically the involved uncertainties in the procedures. In this work it is shown how the measurement uncertainties of an individual calibration can be estimated and how it can be generalized to be valid for others radiation survey meters. (authors)

  17. Calibration of radon-222 detectors using closed circuit radium-226 sources

    International Nuclear Information System (INIS)

    Perna, Allan Felipe Nunes; Paschuk, Sergei Anatolyevich; Correa, Janine Nicolosi; Del Claro, Flavia

    2012-01-01

    This paper presents the results of the calibration of the Radon-222 detectors used by the Laboratories specializing in measuring natural radiation from this gas. The research was conducted in collaboration between UTFPR, CDTN/CNEN, UFRN and IRD/CNEN. During the calibration the detectors were exposed in isolated chambers with radioactive calibrated sources. The calibration procedure was supported with four instant radon monitors AlphaGUARD (SAPHYMO Co.) responsible for radon activity measurements in the experimental chamber. The calibration procedure resulted an equation that relates the number of tracks found in solid-state detector CR-39 (Track-Etch detector) with the concentration of radon in the atmosphere. Obtained results are compatible with previously performed calibration at the National Institute of Radiological Sciences (NIRS, Japan) using high activity levels of radon in air. Present results of calibration give the possibility to expand the calibration curve of CR-39 for medium and low activity levels of radon. (author)

  18. The Architecture Design of Detection and Calibration System for High-voltage Electrical Equipment

    Science.gov (United States)

    Ma, Y.; Lin, Y.; Yang, Y.; Gu, Ch; Yang, F.; Zou, L. D.

    2018-01-01

    With the construction of Material Quality Inspection Center of Shandong electric power company, Electric Power Research Institute takes on more jobs on quality analysis and laboratory calibration for high-voltage electrical equipment, and informationization construction becomes urgent. In the paper we design a consolidated system, which implements the electronic management and online automation process for material sampling, test apparatus detection and field test. In the three jobs we use QR code scanning, online Word editing and electronic signature. These techniques simplify the complex process of warehouse management and testing report transferring, and largely reduce the manual procedure. The construction of the standardized detection information platform realizes the integrated management of high-voltage electrical equipment from their networking, running to periodic detection. According to system operation evaluation, the speed of transferring report is doubled, and querying data is also easier and faster.

  19. The ENEA calibration service for ionising radiations

    International Nuclear Information System (INIS)

    Monteventi, F.; Sermenghi, I.

    1999-01-01

    The report describes all the facilities available at the the service of the ENEA Calibration Service for Ionising Radiations at Bologna (Italy). It gives a detailed description of all equipments qualified for photon fields metrology including the secondary standards and the calibration procedures performed for radiation monitoring devices and dosemeters [it

  20. Sensor Calibration Design Based on D-Optimality Criterion

    Directory of Open Access Journals (Sweden)

    Hajiyev Chingiz

    2016-09-01

    Full Text Available In this study, a procedure for optimal selection of measurement points using the D-optimality criterion to find the best calibration curves of measurement sensors is proposed. The coefficients of calibration curve are evaluated by applying the classical Least Squares Method (LSM. As an example, the problem of optimal selection for standard pressure setters when calibrating a differential pressure sensor is solved. The values obtained from the D-optimum measurement points for calibration of the differential pressure sensor are compared with those from actual experiments. Comparison of the calibration errors corresponding to the D-optimal, A-optimal and Equidistant calibration curves is done.

  1. Calibration of reference KAP-meters at SSDL and cross calibration of clinical KAP-meters

    International Nuclear Information System (INIS)

    Hetland, Per O.; Friberg, Eva G.; Oevreboe, Kirsti M.; Bjerke, Hans H.

    2009-01-01

    In the summer of 2007 the secondary standard dosimetry laboratory (SSDL) in Norway established a calibration service for reference air-kerma product meter (KAP-meter). The air-kerma area product, PKA, is a dosimetric quantity that can be directly related to the patient dose and used for risk assessment associated with different x-ray examinations. The calibration of reference KAP-meters at the SSDL gives important information on parameters influencing the calibration factor for different types of KAP-meters. The use of reference KAP-meters calibrated at the SSDL is an easy and reliable way to calibrate or verify the PKA indicated by the x-ray equipment out in the clinics. Material and methods. Twelve KAP-meters were calibrated at the SSDL by use of the substitution method at five diagnostic radiation qualities (RQRs). Results. The calibration factors varied from 0.94 to 1.18. The energy response of the individual KAP-meters varied by a total of 20% between the different RQRs and the typical chamber transmission factors ranged from 0.78 to 0.91. Discussion. It is important to use a calibrated reference KAP-meter and a harmonised calibration method in the PKA calibration in hospitals. The obtained uncertainty in the PKA readings is comparable with other calibration methods if the information in the calibration certificate is correct used, corrections are made and proper positioning of the KAP-chamber is performed. This will ensure a reliable estimate of the patient dose and a proper optimisation of conventional x-ray examinations and interventional procedures

  2. Time averaging procedure for calculating the mass and energy transfer rates in adiabatic two phase flow

    International Nuclear Information System (INIS)

    Boccaccini, L.V.

    1986-07-01

    To take advantages of the semi-implicit computer models - to solve the two phase flow differential system - a proper averaging procedure is also needed for the source terms. In fact, in some cases, the correlations normally used for the source terms - not time averaged - fail using the theoretical time step that arises from the linear stability analysis used on the right handside. Such a time averaging procedure is developed with reference to the bubbly flow regime. Moreover, the concept of mass that must be exchanged to reach equilibrium from a non-equilibrium state is introduced to limit the mass transfer during a time step. Finally some practical calculations are performed to compare the different correlations for the average mass transfer rate developed in this work. (orig.) [de

  3. Calibration of the indicators of the product KERMa-AREA in radiodiagnostic equipment

    International Nuclear Information System (INIS)

    Ginjaume, M.; Jarvinen, H.; Turak, O.

    2013-01-01

    Currently the IAEA proposes two procedures for the calibration of the determination of the PKA systems integrated in the x-ray equipment. The traditional method is to measure air kerma at a certain distance from the focus through an ionization chamber reference and then the area of the field of radiation in that position. The alternative method is based on measuring directly the product kerma-area using equipment calibrated in this magnitude. In both cases the calibration procedure takes into account the position of the tube on the stretcher of the patient. The objective of this study was to compare the practical application of both calibration procedures as well as assess the accuracy of indication of the PKA in the consoles of the radiology equipment. (Author)

  4. Development of a standard operating procedure for mammography equipment used in calibration of ionized chambers

    International Nuclear Information System (INIS)

    Rodrigues, Yklys Santos; Potiens, Maria da Penha Albuquerque

    2011-01-01

    Mammography is one widely used technique in the detection of breast cancer. In order to optimize the results achieving better images with lower dose rates, a quality assurance programme must be applied to the equipment. Some control tests use ionization chambers to measure air kerma and other quantities. These tests can only be reliable if the ionization chambers used on them are correctly calibrated. In the present work, it was developed a standard operating procedure (SOP) for quality control tests in a commercial mammography equipment installed in the Calibration Laboratory (LCI) at IPEN - Brazilian Institute for energy and nuclear research). Seven tests were performed in the equipment: Tube voltage and exposition time accuracy and reproducibility, linearity and reproducibility of Air kerma and Half Value Layer (HVL). Then, it was made a measurement of the air kerma in the mammography equipment, using a reference ionization chamber with traceability to a primary laboratory in Germany (Physikalisch-Technische Bundesanstalt - PTB), that was later compared with the air kerma measured in an industrial irradiator. This industrial X-ray generator was recently used in the implementation of X-radiation Standards beams, mammography level, following the Standard IEC 61267. The HVL values varied from 0.36 (25kV) to 0.41 mmA1 (35kV), and the measured air kerma rates were between 9.78 and 17.97 mGy/min. (author)

  5. 40 CFR 91.424 - Dilute sampling procedure-CVS calibration.

    Science.gov (United States)

    2010-07-01

    ... radius or American Society of Mechanical Engineers (ASME) flow nozzles are considered equivalent if..., kPa (inches fluid). (iv) A linear least squares fit is performed to generate the calibration... density of exhaust hydrocarbons. (5) The gravimetric mass is subtracted from the CVS measured mass and...

  6. Status of radiation dosimetry in Germany using ionization chamber calibrated in terms of absorbed dose to water

    International Nuclear Information System (INIS)

    Hohlfeld, Klaus; Roos, Martin

    1995-01-01

    In 1984 the PTB as PSDL and the DIN Standard Committee on Radiology (NAR) in close co-operation decided that in Germany the measured absorbed dose to water in a water phantom should replace exposure in the dosimetry for radiation therapy. The PTB has established primary standards of water absorbed dose in the whole range of photon and electron radiation, and international comparisons at the BIPM and with other PSDLs proved agreement within 0.5%. Secondary standards are calibrated in a water phantom under reference conditions in a Co-60 gamma radiation beam at the PTB. Thus, the calibration factor in terms of water absorbed dose, N W , is transferred to the manufacturers of dosimeters, the German Calibration Service and the dosimetry laboratories of the verification authorities. The Verification Law subjects each ionization dosimeter used in the treatment of patients with external photon radiation beams under a type-test at PTB and under a verification procedure, where the calibration factor, N W , must be shown to be within given limits. The absorbed dose determination at the users' level follows the foralism prescribed in the Standard DIN 6800-2 (1995) 'Procedures for Absorbed Dose Determination in Radiology by the Ionization Method'. The concept of this DIN Standard uses exclusively one quantity from the primary standard to the user's instrument eliminating uncertainties and sources of mistakes associated with the conversion of a calibration factor. The concept is simple and clear and covers the whole range of photon and electron radiation. As a means of quality assurance in basic dosimetry the PTB runs a calibration service, up to now on a voluntary basis, which allows the user to compare his dosimetry system against PTB standards using mailed Fricke ampoules, with water absorbed dose as measured and used

  7. Calibration of clinical dosemeters in the IAEA water phantom

    International Nuclear Information System (INIS)

    Caldas, L.V.E.; Albuquerque, M.P.P.

    1994-01-01

    The procedures recommended by the IAEA Code of Practice were applied at the Calibration Laboratory of Sao Paulo in order to provide in the future the clinical dosemeters users with absorbed dose to water calibration factors for Cobalt 60 radiation beams. In this work the clinical dosemeters were calibrated free in air and in water, and the results were compared, using conversion factors. The several tested clinical dosemeters of different manufacturers and models belong to the laboratory and to hospitals. For the measurements in water the IAEA cubic water phantom was used. The dosemeters were all calibrated free in air in terms of air kerma, and the calibration factors in terms of absorbed dose to water were obtained through conversion factors. the same dosemeters were also calibrated into the water phantom. Good agreement was found between the two methods, the differences were always less than 0.5%. The data obtained during this work show that when the dosemeters are used only in Cobalt 60 radiation and the users apply in the hospital routine work the IAEA Code of Practice, the calibration can be performed directly in the water phantom. This procedure provides the useful calibration factors in terms of absorbed dose to water

  8. Calibration of photon and beta ray sources used in brachytherapy. Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories

    International Nuclear Information System (INIS)

    2004-03-01

    It has generally been recognized that international harmonization in radiotherapy dosimetry is essential. Consequently, the IAEA has given much effort to this, for example by publishing a number of reports in the Technical Reports Series (TRS) for external beam dosimetry, most notably TRS-277 and more recently TRS-398. Both of these reports describe in detail the steps to be taken for absorbed dose determination in water and they are often referred to as 'dosimetry protocols'. Similar to TRS-277, it is expected that TRS-398 will be adopted or used as a model by a large number of countries as their national protocol. In 1996, the IAEA established a calibration service for low dose rate (LDR) 137 Cs brachytherapy sources, which is the most widely used source for treatment of gynecological cancer. To further enhance harmonization in brachytherapy dosimetry, the IAEA published in 1999 IAEA-TECDOC-1079 entitled 'Calibration of Brachytherapy Sources. Guidelines on Standardized Procedures for the Calibration of Brachytherapy Sources at Secondary Standard Dosimetry Laboratories (SSDLs) and Hospitals'. The report was well received and was distributed in a large number of copies to the members of the IAEA/WHO network of SSDLs and to medical physicists working with brachytherapy. The present report is an update of the aforementioned TECDOC. Whereas TECDOC-1079 described methods for calibrating brachytherapy sources with photon energies at or above those of 192 Ir, the current report has a wider scope in that it deals with standardization of calibration of all the most commonly used brachytherapy sources, including both photon and beta emitting sources. The latter sources have been in use for a few decades already, but their calibration methods have been unclear. Methods are also described for calibrating sources used in the rapidly growing field of cardiovascular angioplasty. In this application, irradiation of the vessel wall is done in an attempt to prevent restenosis after

  9. [Fundamental aspects for accrediting medical equipment calibration laboratories in Colombia].

    Science.gov (United States)

    Llamosa-Rincón, Luis E; López-Isaza, Giovanni A; Villarreal-Castro, Milton F

    2010-02-01

    Analysing the fundamental methodological aspects which should be considered when drawing up calibration procedure for electro-medical equipment, thereby permitting international standard-based accreditation of electro-medical metrology laboratories in Colombia. NTC-ISO-IEC 17025:2005 and GTC-51-based procedures for calibrating electro-medical equipment were implemented and then used as patterns. The mathematical model for determining the estimated uncertainty value when calibrating electro-medical equipment for accreditation by the Electrical Variable Metrology Laboratory's Electro-medical Equipment Calibration Area accredited in compliance with Superintendence of Industry and Commerce Resolution 25771 May 26th 2009 consists of two equations depending on the case; they are: E = (Ai + sigmaAi) - (Ar + sigmaAr + deltaAr1) and E = (Ai + sigmaAi) - (Ar + sigmaA + deltaAr1). The mathematical modelling implemented for measuring uncertainty in the Universidad Tecnológica de Pereira's Electrical Variable Metrology Laboratory (Electro-medical Equipment Calibration Area) will become a good guide for calibration initiated in other laboratories in Colombia and Latin-America.

  10. Radioactive standards and calibration methods for contamination monitoring instruments

    Energy Technology Data Exchange (ETDEWEB)

    Yoshida, Makoto [Japan Atomic Energy Research Inst., Tokai, Ibaraki (Japan). Tokai Research Establishment

    1997-06-01

    Contamination monitoring in the facilities for handling unsealed radioactive materials is one of the most important procedures for radiation protection as well as radiation dose monitoring. For implementation of the proper contamination monitoring, radiation measuring instruments should not only be suitable to the purpose of monitoring, but also be well calibrated for the objective qualities of measurement. In the calibration of contamination monitoring instruments, quality reference activities need to be used. They are supplied in different such as extended sources, radioactive solutions or radioactive gases. These reference activities must be traceable to the national standards or equivalent standards. On the other hand, the appropriate calibration methods must be applied for each type of contamination monitoring instruments. In this paper, the concepts of calibration for contamination monitoring instruments, reference sources, determination methods of reference quantities and practical calibration methods of contamination monitoring instruments, including the procedures carried out in Japan Atomic Energy Research Institute and some relevant experimental data. (G.K.)

  11. Non-uniformity calibration for MWIR polarization imagery obtained with integrated microgrid polarimeters

    Science.gov (United States)

    Liu, Hai-Zheng; Shi, Ze-Lin; Feng, Bin; Hui, Bin; Zhao, Yao-Hong

    2016-03-01

    Integrating microgrid polarimeters on focal plane array (FPA) of an infrared detector causes non-uniformity of polarization response. In order to reduce the effect of polarization non-uniformity, this paper constructs an experimental setup for capturing raw flat-field images and proposes a procedure for acquiring non-uniform calibration (NUC) matrix and calibrating raw polarization images. The proposed procedure takes the incident radiation as a polarization vector and offers a calibration matrix for each pixel. Both our matrix calibration and two-point calibration are applied to our mid-wavelength infrared (MWIR) polarization imaging system with integrated microgrid polarimeters. Compared with two point calibration, our matrix calibration reduces non-uniformity by 30 40% under condition of flat-field data test with polarization. The ourdoor scene observation experiment indicates that our calibration can effectively reduce polarization non-uniformity and improve the image quality of our MWIR polarization imaging system.

  12. Accurate and Simple Calibration of DLP Projector Systems

    DEFF Research Database (Denmark)

    Wilm, Jakob; Olesen, Oline Vinter; Larsen, Rasmus

    2014-01-01

    does not rely on an initial camera calibration, and so does not carry over the error into projector calibration. A radial interpolation scheme is used to convert features coordinates into projector space, thereby allowing for a very accurate procedure. This allows for highly accurate determination...

  13. Transfer of calibration between length and sweet-spot perception by dynamic touch

    NARCIS (Netherlands)

    Withagen, Rob; Michaels, Claire F.

    2007-01-01

    Calibration is the process that scales perceptual judgment or action to information. An earlier study (Withagen & Michaels, 2004) suggested that perceptual calibration is specific to information-to-perception relations. In the present experiments, the authors tested this hypothesis by asking whether

  14. Calibration Transfer in LIBS and Raman Spectroscopy for Planetary Applications

    Science.gov (United States)

    Dyar, M. D.; Thomas, B. F.; Parente, M.; Gemp, I.; Mullen, T. H.

    2017-12-01

    Planetary scientists rely on spectral libraries and instrument reproducibility to interpret results from missions. Major investments have been made into assembling libraries, but they often naively assume that spectra of single crystals versus powders and from varying instruments will be the same. Calibration transfer (CT) seeks to algorithmically resolve discrepancies among datasets from different instruments or conditions. It offers the ability to align suites of spectra with a small number of common samples, allowing better models to be built with combined data sets. LIBS and Raman data present different challenges for CT. Quantitative geochemical analyses by LIBS spectroscopy are limited by lack of consistency among repeated laser shots and across instruments. Many different factors affect the presence/absence of emission lines and their intensities, such as laser power/plasma temperature, angle of incidence, detector sensitivity/resolution. To overcome these, models in which disparate datasets are projected into a joint low-dimensional subspace where all data can be aligned before quantitative analysis, such as Correlation Analysis for Domain Adaptation (CADA), have proven very effective. They require some overlap between the populations of spectra to be aligned. For example, prediction of SiO2 on 80 samples from two different LIBS labs show errors of ±16-29 wt.% when the training and test sets have no overlap, and ±4.94 wt% SiO2 when CADA is used. Uncorrected Earth-Mars spectral differences are likely to cause errors with the same order of magnitude. As with other types of reflectance spectroscopy, Raman data are plagued by differences among single crystal/powder samples and laser wavelength that affect peak intensities, and by spectral offsets from instruments with varying resolution and wavenumber alignment schemes. These problems persist even within the archetypal RRUFF database. Pre-processing transformation functions such as optimized baseline removal

  15. LANL MTI calibration team experience

    Science.gov (United States)

    Bender, Steven C.; Atkins, William H.; Clodius, William B.; Little, Cynthia K.; Christensen, R. Wynn

    2004-01-01

    The Multispectral Thermal Imager (MTI) was designed as an imaging radiometer with absolute calibration requirements established by Department of Energy (DOE) mission goals. Particular emphasis was given to water surface temperature retrieval using two mid wave and three long wave infrared spectral bands, the fundamental requirement was a surface temperature determination of 1K at the 68% confidence level. For the ten solar reflective bands a one-sigma radiometric performance goal of 3% was established. In order to address these technical challenges a calibration facility was constructed containing newly designed sources that were calibrated at NIST. Additionally, the design of the payload and its onboard calibration system supported post launch maintenance and update of the ground calibration. The on-orbit calibration philosophy also included vicarious techniques using ocean buoys, playas and other instrumented sites; these became increasingly important subsequent to an electrical failure which disabled the onboard calibration system. This paper offers various relevant lessons learned in the eight-year process of reducing to practice the calibration capability required by the scientific mission. The discussion presented will include observations pertinent to operational and procedural issues as well as hardware experiences; the validity of some of the initial assumptions will also be explored.

  16. Experimental Validation of an Efficient Fan-Beam Calibration Procedure for k-Nearest Neighbor Position Estimation in Monolithic Scintillator Detectors

    Science.gov (United States)

    Borghi, Giacomo; Tabacchini, Valerio; Seifert, Stefan; Schaart, Dennis R.

    2015-02-01

    Monolithic scintillator detectors can achieve excellent spatial resolution and coincidence resolving time. However, their practical use for positron emission tomography (PET) and other applications in the medical imaging field is still limited due to drawbacks of the different methods used to estimate the position of interaction. Common statistical methods for example require the collection of an extensive dataset of reference events with a narrow pencil beam aimed at a fine grid of reference positions. Such procedures are time consuming and not straightforwardly implemented in systems composed of many detectors. Here, we experimentally demonstrate for the first time a new calibration procedure for k-nearest neighbor ( k-NN) position estimation that utilizes reference data acquired with a fan beam. The procedure is tested on two detectors consisting of 16 mm ×16 mm ×10 mm and 16 mm ×16 mm ×20 mm monolithic, Ca-codoped LSO:Ce crystals and digital photon counter (DPC) arrays. For both detectors, the spatial resolution and the bias obtained with the new method are found to be practically the same as those obtained with the previously used method based on pencil-beam irradiation, while the calibration time is reduced by a factor of 20. Specifically, a FWHM of 1.1 mm and a FWTM of 2.7 mm were obtained using the fan-beam method with the 10 mm crystal, whereas a FWHM of 1.5 mm and a FWTM of 6 mm were achieved with the 20 mm crystal. Using a fan beam made with a 4.5 MBq 22Na point-source and a tungsten slit collimator with 0.5 mm aperture, the total measurement time needed to acquire the reference dataset was 3 hours for the thinner crystal and 2 hours for the thicker one.

  17. Validation of IEEE P1547.1 Interconnection Test Procedures: ASCO 7000 Soft Load Transfer System

    Energy Technology Data Exchange (ETDEWEB)

    Kroposki, B.; Englebretson, S.; Pink, C.; Daley, J.; Siciliano, R.; Hinton, D.

    2003-09-01

    This report presents the preliminary results of testing the ASCO 7000 Soft Load Transfer System according to IEEE P1547.1 procedures. The ASCO system interconnects synchronous generators with the electric power system and provides monitoring and control for the generator and grid connection through extensive protective functions. The purpose of this testing is to evaluate and give feedback on the contents of IEEE Draft Standard P1547.1 Conformance Tests Procedures for Equipment Interconnecting Distributed Resources With Electric Power Systems.

  18. Wind Tunnel Balance Calibration: Are 1,000,000 Data Points Enough?

    Science.gov (United States)

    Rhew, Ray D.; Parker, Peter A.

    2016-01-01

    Measurement systems are typically calibrated based on standard practices established by a metrology standards laboratory, for example the National Institute for Standards and Technology (NIST), or dictated by an organization's metrology manual. Therefore, the calibration is designed and executed according to an established procedure. However, for many aerodynamic research measurement systems a universally accepted standard, traceable approach does not exist. Therefore, a strategy for how to develop a calibration protocol is left to the developer or user to define based on experience and recommended practice in their respective industry. Wind tunnel balances are one such measurement system. Many different calibration systems, load schedules and procedures have been developed for balances with little consensus on a recommended approach. Especially lacking is guidance the number of calibration data points needed. Regrettably, the number of data points tends to be correlated with the perceived quality of the calibration. Often, the number of data points is associated with ones ability to generate the data rather than by a defined need in support of measurement objectives. Hence the title of the paper was conceived to challenge recent observations in the wind tunnel balance community that shows an ever increasing desire for more data points per calibration absent of guidance to determine when there are enough. This paper presents fundamental concepts and theory to aid in the development of calibration procedures for wind tunnel balances and provides a framework that is generally applicable to the characterization and calibration of other measurement systems. Questions that need to be answered are for example: What constitutes an adequate calibration? How much data are needed in the calibration? How good is the calibration? This paper will assist a practitioner in answering these questions by presenting an underlying theory on how to evaluate a calibration based on

  19. The study of new calibration features in the Harshaw TLD system

    International Nuclear Information System (INIS)

    Luo, L.Z.

    2005-01-01

    Full text: In the Harshaw TLD system, there are three key calibration procedures; the reader, the dosimeter and the algorithm. These procedures must be properly calibrated for the system to achieve the optimum results. For the conventional reader and dosimeter calibration, Harshaw recommends a pre-fade and a post-fade of 24 - 48 hours when calibrating the system. Harshaw also emphasizes that keeping the fade time consistent is important to maintain the quality of the system performance. In recent years, new calibration features have been introduced into the operating system. These new features are auto calibration, auto QC and auto blank, and they give the user the ability to set up the clear-expose-read performed automatically in a sequence for each dosimeter. This saves processing time and keeps the fade time the same. However, since the fade time is near zero, will it affect the TLD system calibration factors? What should the user expect? This paper presents a study of the effect of auto calibration / auto QC to the TLD operation. (author)

  20. Mercury CEM Calibration

    Energy Technology Data Exchange (ETDEWEB)

    John Schabron; Joseph Rovani; Mark Sanderson

    2008-02-29

    Mercury continuous emissions monitoring systems (CEMS) are being implemented in over 800 coal-fired power plant stacks. The power industry desires to conduct at least a full year of monitoring before the formal monitoring and reporting requirement begins on January 1, 2009. It is important for the industry to have available reliable, turnkey equipment from CEM vendors. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The generators are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 requires that calibration be performed with NIST-traceable standards (Federal Register 2007). Traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued an interim traceability protocol for elemental mercury generators (EPA 2007). The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The document is divided into two separate sections. The first deals with the qualification of generators by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the generator models that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma/mass spectrometry performed by NIST in Gaithersburg, MD. The

  1. Calibration of ionization chambers used in LDR brachytherapy

    International Nuclear Information System (INIS)

    Alvarez, Oscar T.B.; Caldas, Linda V.E.

    2005-01-01

    In this work was developed a calibration procedure of well-type ionization chambers used for measurements of I-125, seed type. It was used as a standard an ionization chamber Capintec CRC-15BT, with calibration certificate of the University of Wisconsin. It were calibrated two well-type ionization chambers of Capintec CRC-15R model. A source of I-125 was used in clinical use (18.5 to 7.4 MBq). The results showed that with the application of calibration factors was possible to decrease read deviate from 16% to just 1.0%

  2. Balanced calibration of resonant piezoelectric RL shunts with quasi-static background flexibility correction

    DEFF Research Database (Denmark)

    Høgsberg, Jan Becker; Krenk, Steen

    2015-01-01

    Resonant RL shunt circuits constitute a robust approach to piezoelectric damping, where the performance with respect to damping of flexible structures requires a precise calibration of the corresponding circuit components. The balanced calibration procedure of the present paper is based on equal ...... that the procedure leads to equal modal damping and effective response reduction, even for rather indirect placement of the transducer, provided that the correction for background flexibility is included in the calibration procedure....

  3. Automation is an Effective Way to Improve Quality of Verification (Calibration) of Measuring Instruments

    Science.gov (United States)

    Golobokov, M.; Danilevich, S.

    2018-04-01

    In order to assess calibration reliability and automate such assessment, procedures for data collection and simulation study of thermal imager calibration procedure have been elaborated. The existing calibration techniques do not always provide high reliability. A new method for analyzing the existing calibration techniques and developing new efficient ones has been suggested and tested. A type of software has been studied that allows generating instrument calibration reports automatically, monitoring their proper configuration, processing measurement results and assessing instrument validity. The use of such software allows reducing man-hours spent on finalization of calibration data 2 to 5 times and eliminating a whole set of typical operator errors.

  4. NIST display colorimeter calibration facility

    Science.gov (United States)

    Brown, Steven W.; Ohno, Yoshihiro

    2003-07-01

    A facility has been developed at the National Institute of Standards and Technology (NIST) to provide calibration services for color-measuring instruments to address the need for improving and certifying the measurement uncertainties of this type of instrument. While NIST has active programs in photometry, flat panel display metrology, and color and appearance measurements, these are the first services offered by NIST tailored to color-measuring instruments for displays. An overview of the facility, the calibration approach, and associated uncertainties are presented. Details of a new tunable colorimetric source and the development of new transfer standard instruments are discussed.

  5. Calibration of radiation monitoring instruments

    International Nuclear Information System (INIS)

    1973-01-01

    Radiation protection is dependent on good radiation monitoring, and properly calibrated instruments are essential for this work. Simple procedures for periodically checking and recalibrating different kinds of radiation monitoring instruments are shown in this training film

  6. Calibration of radiation monitoring instruments

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    1974-12-31

    Radiation protection is dependent on good radiation monitoring, and properly calibrated instruments are essential for this work. Simple procedures for periodically checking and recalibrating different kinds of radiation monitoring instruments are shown in this training film

  7. Alternative technique to neutron probe calibration in situ

    International Nuclear Information System (INIS)

    Encarnacao, F.; Carneiro, C.; Dall'Olio, A.

    1990-01-01

    An alternative technique of neutron probe calibration in situ was applied for Podzolic soil. Under field condition, the neutron probe calibration was performed using a special arrangement that prevented the lateral movement of water around the access tube of the neutron probe. During the experiments, successive amounts of water were uniformly infiltrated through the soil profile. Two plots were set to study the effect of the plot dimension on the slope of the calibration curve. The results obtained shown that the amounts of water transferred to the soil profile were significantly correlated to the integrals of count ratio along the soil profile on both plots. In consequence, the slope of calibration curve in field condition was determined. (author)

  8. Traceable calibration for a digital real-time oscilloscope with time interleaving architecture

    Science.gov (United States)

    Kim, Dongju; Lee, Joo-Gwang; Lee, Dong-Joon; Cho, Chihyun

    2018-01-01

    Impairments of analog-to-digital converters (ADCs) used in digital real-time oscilloscopes (DRTO) have caused inevitable signal distortions in measurements. To calibrate these errors with traceability, we propose a novel method that consists of two steps. First, each transfer function of the ADCs is measured using pulse trains from a photodiode calibrated up to 110 GHz. Each data set of the ADCs is superimposed to convert the repetitive pulse to a single pulse to solve the under-sampling problem of the separated data depending on each ADC. Then, the signals of the device under test (DUT) are also separated and superimposed depending on the ADCs, and they are calibrated in the frequency domain based on the measured transfer functions. After a calibration process, the data set is reconverted to the time domain to achieve traceable calibration. To verify our method, we have measured the output of another 70 GHz photodiode with a calibrated DRTO. In terms of results, time-interleaved errors are suppressed by more than 24 dB up to the bandwidth of the DRTO.

  9. The KLOE online calibration system

    International Nuclear Information System (INIS)

    Pasqualucci, E.; Alexander, G.; Aloisio, A.

    2001-01-01

    Based on all the features of the KLOE online software, the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed. A calibration manager process controls the system, implementing the interface to the online system, receiving information from the run control and translating its state transitions to a separate state machine. It acts as a 'calibration run controller' and performs failure recovery when requested by a set of process checkers. The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms. A client library and C, fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an book-like interface. Several calibration processes running in parallel in a distributed, multiplatform environment can fill the same histograms, allowing fast external information check. A monitor thread allow remote browsing for visual inspection. Pre-filtered data are read in non-privileged spy mode from the data acquisition system via the Kloe Integrated Dataflow. The main characteristics of the system are presented

  10. Calibration models for high enthalpy calorimetric probes.

    Science.gov (United States)

    Kannel, A

    1978-07-01

    The accuracy of gas-aspirated liquid-cooled calorimetric probes used for measuring the enthalpy of high-temperature gas streams is studied. The error in the differential temperature measurements caused by internal and external heat transfer interactions is considered and quantified by mathematical models. The analysis suggests calibration methods for the evaluation of dimensionless heat transfer parameters in the models, which then can give a more accurate value for the enthalpy of the sample. Calibration models for four types of calorimeters are applied to results from the literature and from our own experiments: a circular slit calorimeter developed by the author, single-cooling jacket probe, double-cooling jacket probe, and split-flow cooling jacket probe. The results show that the models are useful for describing and correcting the temperature measurements.

  11. The ATLAS Electromagnetic Calorimeter Calibration Workshop

    CERN Multimedia

    Hong Ma; Isabelle Wingerter

    The ATLAS Electromagnetic Calorimeter Calibration Workshop took place at LAPP-Annecy from the 1st to the 3rd of October; 45 people attended the workshop. A detailed program was setup before the workshop. The agenda was organised around very focused presentations where questions were raised to allow arguments to be exchanged and answers to be proposed. The main topics were: Electronics calibration Handling of problematic channels Cluster level corrections for electrons and photons Absolute energy scale Streams for calibration samples Calibration constants processing Learning from commissioning Forty-five people attended the workshop. The workshop was on the whole lively and fruitful. Based on years of experience with test beam analysis and Monte Carlo simulation, and the recent operation of the detector in the commissioning, the methods to calibrate the electromagnetic calorimeter are well known. Some of the procedures are being exercised in the commisssioning, which have demonstrated the c...

  12. Toward a standard line for use in multibeam echo sounder calibration

    Science.gov (United States)

    Weber, Thomas C.; Rice, Glen; Smith, Michael

    2018-06-01

    A procedure is suggested in which a relative calibration for the intensity output of a multibeam echo sounder (MBES) can be performed. This procedure identifies a common survey line (i.e., a standard line), over which acoustic backscatter from the seafloor is collected with multiple MBES systems or by the same system multiple times. A location on the standard line which exhibits temporal stability in its seafloor backscatter response is used to bring the intensity output of the multiple MBES systems to a common reference. This relative calibration procedure has utility for MBES users wishing to generate an aggregate seafloor backscatter mosaic using multiple systems, revisiting an area to detect changes in substrate type, and comparing substrate types in the same general area but with different systems or different system settings. The calibration procedure is demonstrated using three different MBES systems over 3 different years in New Castle, NH, USA.

  13. Marine X-band Weather Radar Data Calibration

    DEFF Research Database (Denmark)

    Thorndahl, Søren Liedtke; Rasmussen, Michael R.

    2012-01-01

    estimates. This paper presents some of the challenges in small marine X-band radar calibration by comparing three calibration procedures for assessing the relationship between radar and rain gauge data. Validation shows similar results for precipitation volumes but more diverse results on peak rain......Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis, and real time control purposes. In these contexts, it is allimportant that the radar data is well calibrated and adjusted in order to obtain valid quantitative precipitation...

  14. Calibration of personal dosimeters: Quantities and terminology

    International Nuclear Information System (INIS)

    Aleinikov, V.E.

    1999-01-01

    The numerical results obtained in the interpretation of individual monitoring of external radiation depend not only on the accurate calibration of the radiation measurement instruments involved, but also on the definition of the quantities in term of which these instruments are calibrated The absence of uniformity in terminology not only makes it difficult to understand properly the scientific and technical literature but can also lead to incorrect interpretation of particular concepts and recommendations. In this paper, brief consideration is given to definition of radiation quantities and terminology used in calibration procedures. (author)

  15. Evaluation of an ASM1 Model Calibration Precedure on a Municipal-Industrial Wastewater Treatment Plant

    DEFF Research Database (Denmark)

    Petersen, Britta; Gernaey, Krist; Henze, Mogens

    2002-01-01

    treatment plant. In the case that was studied it was important to have a detailed description of the process dynamics, since the model was to be used as the basis for optimisation scenarios in a later phase. Therefore, a complete model calibration procedure was applied including: (1) a description......The purpose of the calibrated model determines how to approach a model calibration, e.g. which information is needed and to which level of detail the model should be calibrated. A systematic model calibration procedure was therefore defined and evaluated for a municipal–industrial wastewater...

  16. Improving transferability strategies for debris flow susceptibility assessment: Application to the Saponara and Itala catchments (Messina, Italy)

    KAUST Repository

    Cama, M.

    2017-03-30

    Debris flows can be described as rapid gravity-induced mass movements controlled by topography that are usually triggered as a consequence of storm rainfalls. One of the problems when dealing with debris flow recognition is that the eroded surface is usually very shallow and it can be masked by vegetation or fast weathering as early as one-two years after a landslide has occurred. For this reason, even areas that are highly susceptible to debris flow might suffer of a lack of reliable landslide inventories. However, these inventories are necessary for susceptibility assessment. Model transferability, which is based on calibrating a susceptibility model in a training area in order to predict the distribution of debris flows in a target area, might provide an efficient solution to dealing with this limit. However, when applying a transferability procedure, a key point is the optimal selection of the predictors to be included for calibrating the model in the source area. In this paper, the issue of optimal factor selection is analysed by comparing the predictive performances obtained following three different factor selection criteria. The study includes: i) a test of the similarity between the source and the target areas; ii) the calibration of the susceptibility model in the (training) source area, using different criteria for the selection of the predictors; iii) the validation of the models, both at the source (self-validation, through random partition) and at the target (transferring, through spatial partition) areas. The debris flow susceptibility is evaluated here using binary logistic regression through a R-scripted based procedure.Two separate study areas were selected in the Messina province (southern Italy) in its Ionian (Itala catchment) and Tyrrhenian sides (Saponara catchment), each hit by a severe debris flow event (in 2009 and 2011, respectively).The investigation attested that the best fitting model in the calibration areas resulted poorly performing

  17. Efficiency transfer using the GEANT4 code of CERN for HPGe gamma spectrometry

    International Nuclear Information System (INIS)

    Chagren, S.; Tekaya, M.Ben; Reguigui, N.; Gharbi, F.

    2016-01-01

    In this work we apply the GEANT4 code of CERN to calculate the peak efficiency in High Pure Germanium (HPGe) gamma spectrometry using three different procedures. The first is a direct calculation. The second corresponds to the usual case of efficiency transfer between two different configurations at constant emission energy assuming a reference point detection configuration and the third, a new procedure, consists on the transfer of the peak efficiency between two detection configurations emitting the gamma ray in different energies assuming a “virtual” reference point detection configuration. No pre-optimization of the detector geometrical characteristics was performed before the transfer to test the ability of the efficiency transfer to reduce the effect of the ignorance on their real magnitude on the quality of the transferred efficiency. The obtained and measured efficiencies were found in good agreement for the two investigated methods of efficiency transfer. The obtained agreement proves that Monte Carlo method and especially the GEANT4 code constitute an efficient tool to obtain accurate detection efficiency values. The second investigated efficiency transfer procedure is useful to calibrate the HPGe gamma detector for any emission energy value for a voluminous source using one point source detection efficiency emitting in a different energy as a reference efficiency. The calculations preformed in this work were applied to the measurement exercise of the EUROMET428 project. A measurement exercise where an evaluation of the full energy peak efficiencies in the energy range 60–2000 keV for a typical coaxial p-type HpGe detector and several types of source configuration: point sources located at various distances from the detector and a cylindrical box containing three matrices was performed. - Highlights: • The GEANT4 code of CERN has been used to transfer the peak efficiency from point to points and voluminous detection configurations in HPGe gamma

  18. Standardization in dimensional nanometrology: development of a calibration guideline for Scanning Probe Microscopy

    Science.gov (United States)

    Dziomba, Thorsten; Koenders, Ludger; Wilkening, Günter

    2005-10-01

    The continuing miniaturization in many technologies - among them the optical systems - demands high-resolution measurements with uncertainties in the nanometre-range or even well below. A brief introduction of measurement methods used at the micro- & nanometre scale is therefore given as introduction. While a wide range of these methods are well established for the determination of various physical properties down to the nanometric scale, it is Scanning Probe Microscopy (SPM) that provides a unique direct access to topographic surface features in the size range from atomic diameters to some ten or hundred micrometres. With the increasing use of SPMs as quantitative measurement instruments, the demand for standardized calibration routines also for this type of instruments rises. However, except for a few specially designed set-ups mainly at National Metrology Institutes (e. g. PTB in Germany), measurements made with SPMs usually lack traceability to the metre definition. A number of physical transfer standards have therefore been developed and are already available commercially. While detailed knowledge of the standards' properties is a prerequisite for their practical applicability, the calibration procedure itself deserves careful consideration as well. As there is, up to now, no generally accepted concept how to perform SPM calibrations, guidelines are now being developed on various national and international levels, e. g. VDI/VDE-GMA in Germany and ISO. This papers discusses the draft of an SPM calibration guideline by focusing on several critical practical aspects of SPM calibration. The paper intends to invite the readers to take active part in guideline discussions.

  19. Measurements for the energy calibration of the TANSY neutron detectors

    International Nuclear Information System (INIS)

    Drozdowicz, K.; Hoek, M.; Aronsson, D.

    1990-05-01

    The report describes measurements performed for the energy calibration of the TANSY neutron detectors (two arrays of 16 detectors each one). The calibration procedure determines four calibration parameters for each detector. Results of the calibration measurements are given and test measurements are presented. A relation of the neutron detector calibration parameters to producer's data for the photomulipliers is analysed. Also the tests necessary during normal operation of the TANSY neutron spectrometer are elaborated (passive and active tests). A method how to quickly get the calibration parameters for a spare detector in an array of the neutron detectors is included

  20. Procedures for economic distribution of radionuclides in research facilities

    International Nuclear Information System (INIS)

    Perry, N.A.

    1979-01-01

    A radionuclide accountability system for use in a research facility is described. It can be operated manually or adapted for computer use. All radionuclides are ordered, received, distributed and paid for by the Radiological Control Office who keep complete records of date of order, receipt, calibration use, transfer and/or disposal. Wipe leak tests, specific activity and lot number are also recorded. The procedure provides centralized total accountability records, including financial records, of all radionuclide orders, and the economic advantages of combined purchasing. The use of this system in two medical facilities has resulted in considerable financial savings in the first year of operation. (author)

  1. A user's guide to the MultiMet Sensor Management and Calibration Facility

    OpenAIRE

    Pascal, R.W.; Williams, A.; Ahmed, R.

    1991-01-01

    The report describes the operating instructions and procedures for the MultiMet Sensor Management and Calibration Facility. This includes a description of the Meteological database ME'IDB, and the Sensor Management database which organises the large number of sensors required by the Multimet System. Calibration procedures and policies are also described for the various types of sensors used.

  2. CVM-X: Calibrating contingent values with experimental auction markets

    International Nuclear Information System (INIS)

    Fox, J.A.; Shogren, J.F.; Hayes, D.J.; Kliebenstein, J.B.

    1998-01-01

    We design and implement a method, CVM-X, to calibrate hypothetical survey values using experimental auction markets. We test the procedure using consumer willingness-to-pay for irradiated/nonirradiated meat. Our results show that calibration factors for those who favor the irradiation process (0.67-0.69) are less severe than for those with an initial dislike of the process (0.55-0.59), suggesting that calibration may be commodity specific

  3. Calibration of solar radiation measuring instruments. Final report

    Energy Technology Data Exchange (ETDEWEB)

    Bahm, R J; Nakos, J C

    1979-11-01

    A review of solar radiation measurement of instruments and some types of errors is given; and procedures for calibrating solar radiation measuring instruments are detailed. An appendix contains a description of various agencies who perform calibration of solar instruments and a description of the methods they used at the time this report was prepared. (WHK)

  4. Finite element procedures for coupled linear analysis of heat transfer, fluid and solid mechanics

    Science.gov (United States)

    Sutjahjo, Edhi; Chamis, Christos C.

    1993-01-01

    Coupled finite element formulations for fluid mechanics, heat transfer, and solid mechanics are derived from the conservation laws for energy, mass, and momentum. To model the physics of interactions among the participating disciplines, the linearized equations are coupled by combining domain and boundary coupling procedures. Iterative numerical solution strategy is presented to solve the equations, with the partitioning of temporal discretization implemented.

  5. Studies Concerning the ATLAS IBL Calibration Architecture

    CERN Document Server

    Kretz, Moritz; Kugel, Andreas

    With the commissioning of the Insertable B-Layer (IBL) in 2013 at the ATLAS experiment 12~million additional pixels will be added to the current Pixel Detector. While the idea of employing pairs of VME based Read-Out Driver (ROD) and Back of Crate (BOC) cards in the read-out chain remains unchanged, modifications regarding the IBL calibration procedure were introduced to overcome current hardware limitations. The analysis of calibration histograms will no longer be performed on the RODs, but on an external computing farm that is connected to the RODs via Ethernet. This thesis contributes to the new IBL calibration procedure and presents a concept for a scalable software and hardware architecture. An embedded system targeted to the ROD FPGAs is realized for sending data from the RODs to the fit farm servers and benchmarks are carried out with a Linux based networking stack, as well as a standalone software stack. Furthermore, the histogram fitting algorithm currently being employed on the Pixel Detector RODs i...

  6. Laser Calibration of the ATLAS Tile Calorimeter

    CERN Document Server

    Di Gregorio, Giulia; The ATLAS collaboration

    2017-01-01

    High performance stability of the ATLAS Tile Calorimeter is achieved with a set of calibration procedures. One step of the calibration procedure is based on measurements of the response stability to laser excitation of the PMTs that are used to readout the calorimeter cells. A facility to study in lab the PMT stability response is operating in the PISA-INFN laboratories since 2015. Goals of the tests in lab are to study the time evolution of the PMT response to reproduce and to understand the origin of the response drifts seen with the PMT mounted on the Tile calorimeter in its normal operating during LHC run I and run II. A new statistical approach was developed to measure drift of the absolute gain. This approach was applied to both the ATLAS laser calibration data and to data collected in the Pisa local laboratory. The preliminary results from these two studies are shown.

  7. 30 CFR 90.203 - Certified person; maintenance and calibration.

    Science.gov (United States)

    2010-07-01

    ... COAL MINE SAFETY AND HEALTH MANDATORY HEALTH STANDARDS-COAL MINERS WHO HAVE EVIDENCE OF THE DEVELOPMENT OF PNEUMOCONIOSIS Sampling Procedures § 90.203 Certified person; maintenance and calibration. (a) Approved sampling devices shall be maintained and calibrated by a certified person. (b) To be certified, a...

  8. Step wise, multiple objective calibration of a hydrologic model for a snowmelt dominated basin

    Science.gov (United States)

    Hay, L.E.; Leavesley, G.H.; Clark, M.P.; Markstrom, S.L.; Viger, R.J.; Umemoto, M.

    2006-01-01

    The ability to apply a hydrologic model to large numbers of basins for forecasting purposes requires a quick and effective calibration strategy. This paper presents a step wise, multiple objective, automated procedure for hydrologic model calibration. This procedure includes the sequential calibration of a model's simulation of solar radiation (SR), potential evapotranspiration (PET), water balance, and daily runoff. The procedure uses the Shuffled Complex Evolution global search algorithm to calibrate the U.S. Geological Survey's Precipitation Runoff Modeling System in the Yampa River basin of Colorado. This process assures that intermediate states of the model (SR and PET on a monthly mean basis), as well as the water balance and components of the daily hydrograph are simulated, consistently with measured values.

  9. Comparison between calibration methods of ionization chamber type pencil in greatness P_K_L

    International Nuclear Information System (INIS)

    Macedo, E.M.; Pereira, L.C.S.; Ferreira, M.J.; Navarro, V.C.C.; Garcia, I.F.M.; Pires, E.J.; Navarro, M.V.T.

    2016-01-01

    Calibration of radiation meters is indispensable on Quality Assurance Program in Radiodiagnostic procedures, mainly Computed Tomography. Thus, this study aims evaluate two calibration methods of pencil ionization chambers in terms of Kerma-length Product (P_K_L) (a direct substitution method and an indirect one, through Kerma and length measurements). The results showed a good equivalence, with minimal concordance of 98,5% between calibration factors. About uncertainties, both showed similar results (substitution 2.2% and indirect 2.3%), indicating that the last one is better, due the costs reduction to implant this calibration procedure. (author)

  10. Calibrated Pulse-Thermography Procedure for Inspecting HDPE

    Directory of Open Access Journals (Sweden)

    Mohammed A. Omar

    2008-01-01

    Full Text Available This manuscript discusses the application of a pulse-thermography modality to evaluate the integrity of a high-density polyethylene HDPE joint for delamination, in nonintrusive manner. The inspected HDPE structure is a twin-cup shape, molded through extrusion, and the inspection system comprises a high-intensity, short-duration radiation pulse to excite thermal emission; the text calibrates the experiment settings (pulse duration, and detector sampling rate to accommodate HDPE bulks thermal response. The acquired thermal scans are processed through new contrast computation named “self-referencing”, to investigate the joint tensile strength and further map its adhesion interface in real-time. The proposed system (hardware, software combination performance is assessed through an ultrasound C-scan validation and further benchmarked using a standard pulse phase thermography (PPT routine.

  11. A multi-objective approach to improve SWAT model calibration in alpine catchments

    Science.gov (United States)

    Tuo, Ye; Marcolini, Giorgia; Disse, Markus; Chiogna, Gabriele

    2018-04-01

    Multi-objective hydrological model calibration can represent a valuable solution to reduce model equifinality and parameter uncertainty. The Soil and Water Assessment Tool (SWAT) model is widely applied to investigate water quality and water management issues in alpine catchments. However, the model calibration is generally based on discharge records only, and most of the previous studies have defined a unique set of snow parameters for an entire basin. Only a few studies have considered snow observations to validate model results or have taken into account the possible variability of snow parameters for different subbasins. This work presents and compares three possible calibration approaches. The first two procedures are single-objective calibration procedures, for which all parameters of the SWAT model were calibrated according to river discharge alone. Procedures I and II differ from each other by the assumption used to define snow parameters: The first approach assigned a unique set of snow parameters to the entire basin, whereas the second approach assigned different subbasin-specific sets of snow parameters to each subbasin. The third procedure is a multi-objective calibration, in which we considered snow water equivalent (SWE) information at two different spatial scales (i.e. subbasin and elevation band), in addition to discharge measurements. We tested these approaches in the Upper Adige river basin where a dense network of snow depth measurement stations is available. Only the set of parameters obtained with this multi-objective procedure provided an acceptable prediction of both river discharge and SWE. These findings offer the large community of SWAT users a strategy to improve SWAT modeling in alpine catchments.

  12. High accuracy calibration of a dynamic vapor sorption instrument and determination of the equilibrium humidities using single salts

    DEFF Research Database (Denmark)

    Rörig-Dalgaard, Inge; Svensson, Staffan

    2016-01-01

    We present a procedure for accurately calibrating a dynamic vapor sorption (DVS) instrument using single salts. The procedure accounts for and tailors distinct calibration tests according to the fundamental properties of each salt. Especially relevant properties influencing the calibration are th...

  13. Self-Calibration of CMB Polarimeters

    Science.gov (United States)

    Keating, Brian

    2013-01-01

    Precision measurements of the polarization of the cosmic microwave background (CMB) radiation, especially experiments seeking to detect the odd-parity "B-modes", have far-reaching implications for cosmology. To detect the B-modes generated during inflation the flux response and polarization angle of these experiments must be calibrated to exquisite precision. While suitable flux calibration sources abound, polarization angle calibrators are deficient in many respects. Man-made polarized sources are often not located in the antenna's far-field, have spectral properties that are radically different from the CMB's, are cumbersome to implement and may be inherently unstable over the (long) duration these searches require to detect the faint signature of the inflationary epoch. Astrophysical sources suffer from time, frequency and spatial variability, are not visible from all CMB observatories, and none are understood with sufficient accuracy to calibrate future CMB polarimeters seeking to probe inflationary energy scales of ~1000 TeV. CMB TB and EB modes, expected to identically vanish in the standard cosmological model, can be used to calibrate CMB polarimeters. By enforcing the observed EB and TB power spectra to be consistent with zero, CMB polarimeters can be calibrated to levels not possible with man-made or astrophysical sources. All of this can be accomplished without any loss of observing time using a calibration source which is spectrally identical to the CMB B-modes. The calibration procedure outlined here can be used for any CMB polarimeter.

  14. Dosimetric calibration of solid state detectors with low energy β sources

    International Nuclear Information System (INIS)

    Fidanzio, Andrea; Pia Toni, Maria; Capote, Roberto; Pena, Juan; Pasciuti, Katia; Bovi, Maurizio; Perrone, Franco; Azario, Luigi; Lazzeri, Mauro; Gaudino, Diego; Piermattei, Angelo

    2008-01-01

    A PTW Optidos plastic scintillation and a PTW natural diamond detectors were calibrated in terms of absorbed dose to water with β fields produced by 90 Sr + 90 Y and 85 Kr reference sources. Each source was characterized at the Italian National Metrological Institute - the Istituto Nazionale di Metrologia delle Radiazioni Ionizzanti of ENEA (ENEA-INMRI) - for two different series, 1 and 2, of ISO reference β-particle radiation fields. Beam flattening filters were used for the series 1 β fields to give uniform absorbed dose rates over a large area at a source-to-reference plane distance of 30 cm. The series 2 β fields were produced at source-to-reference plane distance of 10 cm, without the beam flattening filters, in order to obtain higher absorbed dose rates. The reference absorbed dose rate values were directly determined by the Italian national standard for β-particle dosimetry (a PTW extrapolation ionization chamber) for the series 1 β fields and by a calibrated transfer standard chamber, (a Capintec thin fixed-volume parallel plate ionization chamber) for the series 2 β fields. Finally the two solid state detectors were calibrated in terms of absorbed dose to water with the series 2 β field. The expanded uncertainties of the calibration coefficients obtained for the plastic scintillation dosimeter were 10% and 12% (2SD) for the 90 Sr + 90 Y and the 85 Kr sources, respectively. The expanded uncertainties obtained for the diamond dosimeter were 10% (2SD) and 16% (2SD) for the 90 Sr + 90 Y and the 85 Kr sources, respectively. The good results obtained with the 90 Sr + 90 Y and the 85 Kr β sources encourage to implement this procedure to calibrate this type of detectors at shorter distances and with other β sources of interest in brachytherapy, for example the 106 Ru source

  15. Innovative methodology for intercomparison of radionuclide calibrators using short half-life in situ prepared radioactive sources

    International Nuclear Information System (INIS)

    Oliveira, P. A.; Santos, J. A. M.

    2014-01-01

    Purpose: An original radionuclide calibrator method for activity determination is presented. The method could be used for intercomparison surveys for short half-life radioactive sources used in Nuclear Medicine, such as 99m Tc or most positron emission tomography radiopharmaceuticals. Methods: By evaluation of the resulting net optical density (netOD) using a standardized scanning method of irradiated Gafchromic XRQA2 film, a comparison of the netOD measurement with a previously determined calibration curve can be made and the difference between the tested radionuclide calibrator and a radionuclide calibrator used as reference device can be calculated. To estimate the total expected measurement uncertainties, a careful analysis of the methodology, for the case of 99m Tc, was performed: reproducibility determination, scanning conditions, and possible fadeout effects. Since every factor of the activity measurement procedure can influence the final result, the method also evaluates correct syringe positioning inside the radionuclide calibrator. Results: As an alternative to using a calibrated source sent to the surveyed site, which requires a relatively long half-life of the nuclide, or sending a portable calibrated radionuclide calibrator, the proposed method uses a source preparedin situ. An indirect activity determination is achieved by the irradiation of a radiochromic film using 99m Tc under strictly controlled conditions, and cumulated activity calculation from the initial activity and total irradiation time. The irradiated Gafchromic film and the irradiator, without the source, can then be sent to a National Metrology Institute for evaluation of the results. Conclusions: The methodology described in this paper showed to have a good potential for accurate (3%) radionuclide calibrators intercomparison studies for 99m Tc between Nuclear Medicine centers without source transfer and can easily be adapted to other short half-life radionuclides

  16. ONE-STEP AND TWO-STEP CALIBRATION OF A PORTABLE PANORAMIC IMAGE MAPPING SYSTEM

    Directory of Open Access Journals (Sweden)

    P.-C. Wang

    2012-07-01

    Full Text Available A Portable Panoramic Image Mapping System (PPIMS is proposed for rapid acquisition of three-dimensional spatial information. By considering the convenience of use, cost, weight of equipment, precision, and power supply, the designed PPIMS is equipped with 6 circularly arranged cameras to capture panoramic images and a GPS receiver for positioning. The motivation for this design is to develop a hand-held Mobile Mapping System (MMS for some difficult accessing areas by vehicle MMS, such as rugged terrains, forest areas, heavily damaged disaster areas, and crowed places etc. This PPIMS is in fact a GPS assisted close-range photogrammetric system. Compared with the traditional close-range photogrammetry, PPIMS can reduce the need of ground control points significantly. Under the condition of knowing the relative geometric relationships of the equipped sensors, the elements of exterior orientation of each captured image can be solved. However, the procedure of a system calibration should be done accurately to determine the relative geometric relationships of multi-cameras and the GPS antenna center, before the PPIMS can be applied for geo-referenced mapping. In this paper, both of one-step and two-step calibration procedures for PPIMS are performed to determine the lever-arm offsets and boresight angles among cameras and GPS. The performance of the one-step and two-step calibration is evaluated through the analysis of the experimental results. The comparison between these two calibration procedures was also conducted. The two-step calibration method outperforms the one-step calibration method in terms of calibration accuracy and operation convenience. We expect that the proposed two-step calibration procedure can also be applied to other platform-based MMSs.

  17. Calibrating a combined energy systems analysis and controller design method with empirical data

    International Nuclear Information System (INIS)

    Murphy, Gavin Bruce; Counsell, John; Allison, John; Brindley, Joseph

    2013-01-01

    The drive towards low carbon constructions has seen buildings increasingly utilise many different energy systems simultaneously to control the human comfort of the indoor environment; such as ventilation with heat recovery, various heating solutions and applications of renewable energy. This paper describes a dynamic modelling and simulation method (IDEAS – Inverse Dynamics based Energy Assessment and Simulation) for analysing the energy utilisation of a building and its complex servicing systems. The IDEAS case study presented in this paper is based upon small perturbation theory and can be used for the analysis of the performance of complex energy systems and also for the design of smart control systems. This paper presents a process of how any dynamic model can be calibrated against a more empirical based data model, in this case the UK Government's SAP (Standard Assessment Procedure). The research targets of this work are building simulation experts for analysing the energy use of a building and also control engineers to assist in the design of smart control systems for dwellings. The calibration process presented is transferable and has applications for simulation experts to assist in calibrating any dynamic building simulation method with an empirical based method. - Highlights: • Presentation of an energy systems analysis method for assessing the energy utilisation of buildings and their complex servicing systems. • An inverse dynamics based controller design method is detailed. • Method of how a dynamic model can be calibrated with an empirical based model

  18. CryoSat SIRAL Calibration and Performance

    Science.gov (United States)

    Fornari, Marco; Scagliola, Michele; Tagliani, Nicolas; Parrinello, Tommaso

    2013-04-01

    The main payload of CryoSat is a Ku band pulse-width limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for azimuth processing. This allows to reach an along track resolution of about 250 meters which is a significant improvement over traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed, including both an internal and external calibration. The internal calibration monitors the instrument impulse response and the transfer function, like traditional altimeters. In addition to that, the interferometer requires a special calibration developed ad hoc for SIRAL. The external calibration is performed with the use of a ground transponder, located in Svalbard, which receives SIRAL signal and sends the echo back to the satellite. Internal calibration data are processed on ground by the CryoSat Instrument Processing Facility (IPF1) and then applied to the science data. By April 2013, almost 3 years of calibration data will be available, which will be shown in this poster. The external calibration (transponder) data are processed and analyzed independently from the operational chain. The use of an external transponder has been very useful to determine instrument performance and for the tuning of the on-ground processor. This poster presents the transponder results in terms of range noise and datation error.

  19. Calibration Laboratory of the Paul Scherrer Institute

    International Nuclear Information System (INIS)

    Gmuer, K.; Wernli, C.

    1994-01-01

    Calibration and working checks of radiation protection instruments are carried out at the Calibration Laboratory of the Paul Scherrer Institute. In view of the new radiation protection regulation, the calibration laboratory received an official federal status. The accreditation procedure in cooperation with the Federal Office of Metrology enabled a critical review of the techniques and methods applied. Specifically, personal responsibilities, time intervals for recalibration of standard instruments, maximum permissible errors of verification, traceability and accuracy of the standard instruments, form and content of the certificates were defined, and the traceability of the standards and quality assurance were reconsidered. (orig.) [de

  20. A NEW AUTOMATIC SYSTEM CALIBRATION OF MULTI-CAMERAS AND LIDAR SENSORS

    Directory of Open Access Journals (Sweden)

    M. Hassanein

    2016-06-01

    Full Text Available In the last few years, multi-cameras and LIDAR systems draw the attention of the mapping community. They have been deployed on different mobile mapping platforms. The different uses of these platforms, especially the UAVs, offered new applications and developments which require fast and accurate results. The successful calibration of such systems is a key factor to achieve accurate results and for the successful processing of the system measurements especially with the different types of measurements provided by the LIDAR and the cameras. The system calibration aims to estimate the geometric relationships between the different system components. A number of applications require the systems be ready for operation in a short time especially for disasters monitoring applications. Also, many of the present system calibration techniques are constrained with the need of special arrangements in labs for the calibration procedures. In this paper, a new technique for calibration of integrated LIDAR and multi-cameras systems is presented. The new proposed technique offers a calibration solution that overcomes the need for special labs for standard calibration procedures. In the proposed technique, 3D reconstruction of automatically detected and matched image points is used to generate a sparse images-driven point cloud then, a registration between the LIDAR generated 3D point cloud and the images-driven 3D point takes place to estimate the geometric relationships between the cameras and the LIDAR.. In the presented technique a simple 3D artificial target is used to simplify the lab requirements for the calibration procedure. The used target is composed of three intersected plates. The choice of such target geometry was to ensure enough conditions for the convergence of registration between the constructed 3D point clouds from the two systems. The achieved results of the proposed approach prove its ability to provide an adequate and fully automated

  1. Improvement of the calibration technique of clinical dosemeters

    International Nuclear Information System (INIS)

    Ehlin Caldas, L.V.

    1988-08-01

    Clinical dosemeters constituted of ionization chambers connected to electrometers are usually calibrated as whole systems in appropriate radiation fields against secondary standard dosemeters in calibration laboratories. This work reports on a technique of component calibration procedures separately for chambers and electrometers applied in the calibration laboratory of IPEN-CNEN, Brazil. For electrometer calibration, redundancy was established by using a standard capacitor of 1000pF (General Radio, USA) and a standard current source based on air ionization with Sr 90 (PTW, Germany). The results from both methods applied to several electrometers of clinical dosemeters agreed within 0.4%. The calibration factors for the respective chambers were determined by intercomparing their response to the response of a certified calibrated chamber in a Co 60 calibration beam using a Keithley electrometer type 617. Overall calibration factors compared with the product of the respective component calibration factors for the tested dosemeters showed an agreement better than 0.7%. This deviation has to be considered with regard to an uncertainty of 2.5% in routine calibration of clinical dosemeters. Calibration by components permits to calibrate ionization chambers one at a time for those hospitals who have several ionization chambers but only one electrometer (small hospitals, hospitals in developing countries). 6 refs, 2 figs, 2 tabs

  2. Calibration of kV measurers with the practical peak voltage (IEC 1676)

    International Nuclear Information System (INIS)

    Becker, Paulo H.B.; Peres, Marcos A.L.; Ludwig, Jaime L.; Chernicharo, Carlos C.

    2002-01-01

    The IEC 1676 standard introduces a new quantity for the measurements of the high voltages applied to the X ray tubes used for diagnosis, the 'Practical Peak Voltage' (PPV). In order to start the introduction of this new quantity in Brazil the National Laboratory for Metrology of Ionizing Radiation has developed a procedure for calibrating measuring instruments in this quantity. This procedure is based in the same set up used for the calibration of the conventional kVp, which consists in a high voltage divider (Dynalyser III from Radcal Corporation), a fast analogue to digital conversion board and a data acquisition software. In order to evaluate this procedure a commercial kVp measure instrument that is able to measure PPV (Universal Diavolt from PTW) was calibrated and the results compared. This work presents a summary of the procedure developed and the results obtained with the comparison. (author)

  3. Determination of calibration equations by means of the generalized least squares method

    International Nuclear Information System (INIS)

    Zijp, W.L.

    1984-12-01

    For the determination of two-dimensional calibration curves (e.g. in tank calibration procedures) or of three dimensional calibration equations (e.g. for the calibration of NDA equipment for enrichment measurements) one performs measurements under well chosen conditions, where all observables of interest (inclusive the values of the standard material) are subject to measurement uncertainties. Moreover correlations in several measurements may occur. This document describes the mathematical-statistical approach to determine the values of the model parameters and their covariance matrix, which fit best to the mathematical model for the calibration equation. The formulae are based on the method of generalized least squares where the term generalized implies that non-linear equations in the unknown parameters and also covariance matrices of the measurement data of the calibration can be taken into account. In the general case an iteration procedure is required. No iteration is required when the model is linear in the parameters and the covariance matrices for the measurements of co-ordinates of the calibration points are proportional to each other

  4. Efficiency transfer using the GEANT4 code of CERN for HPGe gamma spectrometry.

    Science.gov (United States)

    Chagren, S; Tekaya, M Ben; Reguigui, N; Gharbi, F

    2016-01-01

    In this work we apply the GEANT4 code of CERN to calculate the peak efficiency in High Pure Germanium (HPGe) gamma spectrometry using three different procedures. The first is a direct calculation. The second corresponds to the usual case of efficiency transfer between two different configurations at constant emission energy assuming a reference point detection configuration and the third, a new procedure, consists on the transfer of the peak efficiency between two detection configurations emitting the gamma ray in different energies assuming a "virtual" reference point detection configuration. No pre-optimization of the detector geometrical characteristics was performed before the transfer to test the ability of the efficiency transfer to reduce the effect of the ignorance on their real magnitude on the quality of the transferred efficiency. The obtained and measured efficiencies were found in good agreement for the two investigated methods of efficiency transfer. The obtained agreement proves that Monte Carlo method and especially the GEANT4 code constitute an efficient tool to obtain accurate detection efficiency values. The second investigated efficiency transfer procedure is useful to calibrate the HPGe gamma detector for any emission energy value for a voluminous source using one point source detection efficiency emitting in a different energy as a reference efficiency. The calculations preformed in this work were applied to the measurement exercise of the EUROMET428 project. A measurement exercise where an evaluation of the full energy peak efficiencies in the energy range 60-2000 keV for a typical coaxial p-type HpGe detector and several types of source configuration: point sources located at various distances from the detector and a cylindrical box containing three matrices was performed. Copyright © 2015 Elsevier Ltd. All rights reserved.

  5. Standard practice for torque calibration of testing machines and devices

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2009-01-01

    1.1 This practice covers procedures and requirements for the calibration of torque for static and quasi-static torque capable testing machines or devices. These may, or may not, have torque indicating systems and include those devices used for the calibration of hand torque tools. Testing machines may be calibrated by one of the three following methods or combination thereof: 1.1.1 Use of standard weights and lever arms. 1.1.2 Use of elastic torque measuring devices. 1.1.3 Use of elastic force measuring devices and lever arms. 1.1.4 Any of the methods require a specific uncertainty of measurement and a traceability derived from national standards of mass and length. 1.2 The procedures of 1.1.1, 1.1.2, and 1.1.3 apply to the calibration of the torque-indicating systems associated with the testing machine, such as a scale, dial, marked or unmarked recorder chart, digital display, etc. In all cases the buyer/owner/user must designate the torque-indicating system(s) to be calibrated and included in the repor...

  6. Expert Performance Transfer: Making Knowledge Transfer Count

    International Nuclear Information System (INIS)

    Turner, C.L.; Braudt, T.E.

    2011-01-01

    'Knowledge Transfer' is a high-priority imperative as the nuclear industry faces the combined effects of an aging workforce and economic pressures to do more with less. Knowledge Transfer is only a part of the solution to these challenges, however. The more compelling and immediate need faced by industry is Accomplishment Transfer, or the transference of the applied knowledge necessary to assure optimal performance transfer from experienced, high-performing staff to inexperienced staff. A great deal of industry knowledge and required performance information has been documented in the form of procedures. Often under-appreciated either as knowledge stores or as drivers of human performance, procedures, coupled with tightly-focused and effective training, are arguably the most effective influences on human and plant performance. (author)

  7. Multiple height calibration artefact for 3D microscopy

    DEFF Research Database (Denmark)

    De Chiffre, Leonardo; Carli, Lorenzo; Eriksen, Rasmus Solmer

    2011-01-01

    A novel artefact for calibration of the height in 3D microscopy is presented. The artefact comprises three steps having a common vertical axis, which allows z-coordinate calibration at different magnifications without requiring repositioning. The artefact is suitable for transferring traceability...... to 3D techniques at the micrometer and nanometer scale, e.g. 3D SEM, confocal microscopes etc. Two different series of samples were fabricated using EDM with three steps of 2–5–7μm, and 20–50–70μm, respectively, from a 3mm diameter carbide wire. The artefact steps were calibrated on a stylus instrument...

  8. 3D-radiative transfer in terrestrial atmosphere: An efficient parallel numerical procedure

    Science.gov (United States)

    Bass, L. P.; Germogenova, T. A.; Nikolaeva, O. V.; Kokhanovsky, A. A.; Kuznetsov, V. S.

    2003-04-01

    Light propagation and scattering in terrestrial atmosphere is usually studied in the framework of the 1D radiative transfer theory [1]. However, in reality particles (e.g., ice crystals, solid and liquid aerosols, cloud droplets) are randomly distributed in 3D space. In particular, their concentrations vary both in vertical and horizontal directions. Therefore, 3D effects influence modern cloud and aerosol retrieval procedures, which are currently based on the 1D radiative transfer theory. It should be pointed out that the standard radiative transfer equation allows to study these more complex situations as well [2]. In recent year the parallel version of the 2D and 3D RADUGA code has been developed. This version is successfully used in gammas and neutrons transport problems [3]. Applications of this code to radiative transfer in atmosphere problems are contained in [4]. Possibilities of code RADUGA are presented in [5]. The RADUGA code system is an universal solver of radiative transfer problems for complicated models, including 2D and 3D aerosol and cloud fields with arbitrary scattering anisotropy, light absorption, inhomogeneous underlying surface and topography. Both delta type and distributed light sources can be accounted for in the framework of the algorithm developed. The accurate numerical procedure is based on the new discrete ordinate SWDD scheme [6]. The algorithm is specifically designed for parallel supercomputers. The version RADUGA 5.1(P) can run on MBC1000M [7] (768 processors with 10 Gb of hard disc memory for each processor). The peak productivity is equal 1 Tfl. Corresponding scalar version RADUGA 5.1 is working on PC. As a first example of application of the algorithm developed, we have studied the shadowing effects of clouds on neighboring cloudless atmosphere, depending on the cloud optical thickness, surface albedo, and illumination conditions. This is of importance for modern satellite aerosol retrieval algorithms development. [1] Sobolev

  9. Nimbus-7 Earth radiation budget calibration history. Part 2: The Earth flux channels

    Science.gov (United States)

    Kyle, H. Lee; Hucek, Douglas Richard R.; Ardanuy, Philip E.; Hickey, John R.; Maschhoff, Robert H.; Penn, Lanning M.; Groveman, Brian S.; Vallette, Brenda J.

    1994-01-01

    Nine years (November 1978 to October 1987) of Nimbus-7 Earth radiation budget (ERB) products have shown that the global annual mean emitted longwave, absorbed shortwave, and net radiation were constant to within about + 0.5 W/sq m. Further, most of the small annual variations in the emitted longwave have been shown to be real. To obtain this measurement accuracy, the wide-field-of-view (WFOV) Earth-viewing channels 12 (0.2 to over 50 micrometers), 13 (0.2 to 3.8 micrometers), and 14 (0.7 to 2.8 micrometers) have been characterized in their satellite environment to account for signal variations not considered in the prelaunch calibration equations. Calibration adjustments have been derived for (1) extraterrestrial radiation incident on the detectors, (2) long-term degradation of the sensors, and (3) thermal perturbations within the ERB instrument. The first item is important in all the channels; the second, mainly in channels 13 and 14, and the third, only in channels 13 and 14. The Sun is used as a stable calibration source to monitor the long-term degradation of the various channels. Channel 12, which is reasonably stable to both thermal perturbations and sensor degradation, is used as a reference and calibration transfer agent for the drifting sensitivities of the filtered channels 13 and 14. Redundant calibration procedures were utilized. Laboratory studies complemented analyses of the satellite data. Two nearly independent models were derived to account for the thermal perturbations in channels 13 and 14. The global annual mean terrestrial shortwave and longwave signals proved stable enough to act as secondary calibration sources. Instantaneous measurements may still, at times, be in error by as much as a few Wm(exp -2), but the long-term averages are stable to within a fraction of a Wm(exp -2).

  10. INR calibration of Owren-type prothrombin time based on the relationship between PT% and INR utilizing normal plasma samples.

    Science.gov (United States)

    Lindahl, Tomas L; Egberg, Nils; Hillarp, Andreas; Ødegaard, Ole R; Edlund, Bror; Svensson, Jan; Sandset, Per M; Rånby, Mats

    2004-06-01

    Prothrombin time (PT) is clinically important and is used to monitor oral anticoagulant therapy. To obtain PT results in international normalized ratio (INR), the current standardization procedure is complex and involves reference reagents. The PT of diluted plasma samples can be determined with a combined thromboplastin (the Owren-type procedure), but not necessarily with a plain thromboplastin (the Quick-type procedure). Owren-type PT procedures can therefore, as an alternative to the INR calibration, be calibrated with diluted normal plasma to give PT results in percent of normal PT activity (PT%). The present study explored if a plasma-based calibration of an Owren-type PT procedure can be used to obtain results in INR. The approach was to establish a relationship between PT% and INR by multi-center analysis of 365 samples from healthy individuals and patients on warfarin treatment. INR values were obtained by manual Quick-type reference procedure and PT% values by various automated Owren-type procedures. A relationship INR = (1/PT% + 0.018)/0.028 was found. A calibration procedure, based on the relationship, was investigated. Calibrators were the median PT of 21 normal plasma at dilutions representing 100%, 50%, 25%, 12.5% and 6.25% of normal PT activity. These were assigned INR values of 1.00, 1.36, 2.07, 3.05 and 6.36. Calibration of various Owren-type assays was repeatedly performed by 5 expert laboratories during 3 consecutive years. The INR values of certain lyophilised or frozen control plasmas were determined. The frozen control plasmas had externally assigned INR values according to WHO guide-lines. Within the laboratory, CV was typically below 3%. No appreciable difference among the results of the different laboratories or the three assay occasions was found. Externally assigned and INR values were essentially identical to those found. These and other results indicated that the calibration procedure was reproducible, precise and accurate. Thus, an

  11. Research on orbit prediction for solar-based calibration proper satellite

    Science.gov (United States)

    Chen, Xuan; Qi, Wenwen; Xu, Peng

    2018-03-01

    Utilizing the mathematical model of the orbit mechanics, the orbit prediction is to forecast the space target's orbit information of a certain time based on the orbit of the initial moment. The proper satellite radiometric calibration and calibration orbit prediction process are introduced briefly. On the basis of the research of the calibration space position design method and the radiative transfer model, an orbit prediction method for proper satellite radiometric calibration is proposed to select the appropriate calibration arc for the remote sensor and to predict the orbit information of the proper satellite and the remote sensor. By analyzing the orbit constraint of the proper satellite calibration, the GF-1solar synchronous orbit is chose as the proper satellite orbit in order to simulate the calibration visible durance for different satellites to be calibrated. The results of simulation and analysis provide the basis for the improvement of the radiometric calibration accuracy of the satellite remote sensor, which lays the foundation for the high precision and high frequency radiometric calibration.

  12. Temperature calibration procedure for thin film substrates for thermo-ellipsometric analysis using melting point standards

    International Nuclear Information System (INIS)

    Kappert, Emiel J.; Raaijmakers, Michiel J.T.; Ogieglo, Wojciech; Nijmeijer, Arian; Huiskes, Cindy; Benes, Nieck E.

    2015-01-01

    Highlights: • Facile temperature calibration method for thermo-ellipsometric analysis. • The melting point of thin films of indium, lead, zinc, and water can be detected by ellipsometry. • In-situ calibration of ellipsometry hot stage, without using any external equipment. • High-accuracy temperature calibration (±1.3 °C). - Abstract: Precise and accurate temperature control is pertinent to studying thermally activated processes in thin films. Here, we present a calibration method for the substrate–film interface temperature using spectroscopic ellipsometry. The method is adapted from temperature calibration methods that are well developed for thermogravimetric analysis and differential scanning calorimetry instruments, and is based on probing a transition temperature. Indium, lead, and zinc could be spread on a substrate, and the phase transition of these metals could be detected by a change in the Ψ signal of the ellipsometer. For water, the phase transition could be detected by a loss of signal intensity as a result of light scattering by the ice crystals. The combined approach allowed for construction of a linear calibration curve with an accuracy of 1.3 °C or lower over the full temperature range

  13. On the efficacy of using the transfer-controlled procedure during periods of STP processor overloads in SS7 networks

    Science.gov (United States)

    Rumsewicz, Michael

    1994-04-01

    In this paper, we examine call completion performance, rather than message throughput, in a Common Channel Signaling network in which the processing resources, and not transmission resources, of a Signaling Transfer Point (STP) are overloaded. Specifically, we perform a transient analysis, via simulation, of a network consisting of a single Central Processor-based STP connecting many local exchanges. We consider the efficacy of using the Transfer Controlled (TFC) procedure when the network call attempt rate exceeds the processing capability of the STP. We find the following: (1) the success of the control depends critically on the rate at which TFC's are sent; (2) use of the TFC procedure in theevent of processor overload can provide reasonable call completion rates.

  14. Calibrating EEG-based motor imagery brain-computer interface from passive movement.

    Science.gov (United States)

    Ang, Kai Keng; Guan, Cuntai; Wang, Chuanchu; Phua, Kok Soon; Tan, Adrian Hock Guan; Chin, Zheng Yang

    2011-01-01

    EEG data from performing motor imagery are usually collected to calibrate a subject-specific model for classifying the EEG data during the evaluation phase of motor imagery Brain-Computer Interface (BCI). However, there is no direct objective measure to determine if a subject is performing motor imagery correctly for proper calibration. Studies have shown that passive movement, which is directly observable, induces Event-Related Synchronization patterns that are similar to those induced from motor imagery. Hence, this paper investigates the feasibility of calibrating EEG-based motor imagery BCI from passive movement. EEG data of 12 healthy subjects were collected during motor imagery and passive movement of the hand by a haptic knob robot. The calibration models using the Filter Bank Common Spatial Pattern algorithm on the EEG data from motor imagery were compared against using the EEG data from passive movement. The performances were compared based on the 10×10-fold cross-validation accuracies of the calibration data, and off-line session-to-session transfer kappa values to other sessions of motor imagery performed on another day. The results showed that the calibration performed using passive movement yielded higher model accuracy and off-line session-to-session transfer (73.6% and 0.354) than the calibration performed using motor imagery (71.3% and 0.311), and no significant differences were observed between the two groups (p=0.20, 0.23). Hence, this study shows that it is feasible to calibrate EEG-based motor imagery BCI from passive movement.

  15. Calibration and validation of a model describing complete autotrophic nitrogen removal in a granular SBR system

    DEFF Research Database (Denmark)

    Vangsgaard, Anna Katrine; Mutlu, Ayten Gizem; Gernaey, Krist

    2013-01-01

    BACKGROUND: A validated model describing the nitritation-anammox process in a granular sequencing batch reactor (SBR) system is an important tool for: a) design of future experiments and b) prediction of process performance during optimization, while applying process control, or during system scale......-up. RESULTS: A model was calibrated using a step-wise procedure customized for the specific needs of the system. The important steps in the procedure were initialization, steady-state and dynamic calibration, and validation. A fast and effective initialization approach was developed to approximate pseudo...... screening of the parameter space proposed by Sin et al. (2008) - to find the best fit of the model to dynamic data. Finally, the calibrated model was validated with an independent data set. CONCLUSION: The presented calibration procedure is the first customized procedure for this type of system...

  16. Computerized 50 liter volume calibration system

    International Nuclear Information System (INIS)

    Proffitt, T.H.

    1990-01-01

    A system has been designed for the Savannah River Site that will be used to calibrate product shipping containers. For accountability purposes, it is necessary that these containers be calibrated to a very high precision. The Computerized 50 Liter Volume Calibration System (CVCS), which is based on the Ideal Gas Law (IGL), will use reference volumes with precision of no less ±0.03%, and helium to calibrate the containers to have a total error of no greater than ±0.10%. A statistical interpretation of the system has given a theoretical total calculated error of ±0.08%. Tests with the system will be performed once fabrication is complete to experimentally verify the calculated error. Since the total error was calculated using the worst case scenario, the actual error should be significantly less than the calculated value. The computer controlled, totally automated system is traceable to the National Institute of Standards and Technology. The design, calibration procedure, and statistical interpretation of the system will be discussed. 1 ref

  17. Node-to-node field calibration of wireless distributed air pollution sensor network.

    Science.gov (United States)

    Kizel, Fadi; Etzion, Yael; Shafran-Nathan, Rakefet; Levy, Ilan; Fishbain, Barak; Bartonova, Alena; Broday, David M

    2018-02-01

    Low-cost air quality sensors offer high-resolution spatiotemporal measurements that can be used for air resources management and exposure estimation. Yet, such sensors require frequent calibration to provide reliable data, since even after a laboratory calibration they might not report correct values when they are deployed in the field, due to interference with other pollutants, as a result of sensitivity to environmental conditions and due to sensor aging and drift. Field calibration has been suggested as a means for overcoming these limitations, with the common strategy involving periodical collocations of the sensors at an air quality monitoring station. However, the cost and complexity involved in relocating numerous sensor nodes back and forth, and the loss of data during the repeated calibration periods make this strategy inefficient. This work examines an alternative approach, a node-to-node (N2N) calibration, where only one sensor in each chain is directly calibrated against the reference measurements and the rest of the sensors are calibrated sequentially one against the other while they are deployed and collocated in pairs. The calibration can be performed multiple times as a routine procedure. This procedure minimizes the total number of sensor relocations, and enables calibration while simultaneously collecting data at the deployment sites. We studied N2N chain calibration and the propagation of the calibration error analytically, computationally and experimentally. The in-situ N2N calibration is shown to be generic and applicable for different pollutants, sensing technologies, sensor platforms, chain lengths, and sensor order within the chain. In particular, we show that chain calibration of three nodes, each calibrated for a week, propagate calibration errors that are similar to those found in direct field calibration. Hence, N2N calibration is shown to be suitable for calibration of distributed sensor networks. Copyright © 2017 Elsevier Ltd. All

  18. Calibration of dose meters used in radiotherapy

    International Nuclear Information System (INIS)

    1979-01-01

    This manual is a practical guide, not a comprehensive textbook, to the instrumentation and procedures necessary to calibrate a radiation dose meter used in clinical practice against a secondary standard dose meter

  19. Self-calibrating interferometer

    International Nuclear Information System (INIS)

    Nussmeier, T.A.

    1982-01-01

    A self-calibrating interferometer is disclosed which forms therein a pair of Michelson interferometers with one beam length of each Michelson interferometer being controlled by a common phase shifter. The transfer function measured from the phase shifter to either of a pair of detectors is sinusoidal with a full cycle for each half wavelength of phase shifter travel. The phase difference between these two sinusoidal detector outputs represents the optical phase difference between a path of known distance and a path of unknown distance

  20. 3D-Calibration for IMU of the Strapdown Inertial Navigation Systems

    Directory of Open Access Journals (Sweden)

    Avrutov V.V.

    2017-01-01

    Full Text Available A new calibration method for Inertial Measurement Unit (IMU of Strapdown Iner-tial Navigation Systems was presented. IMU has been composed of accelerometers, gyroscopes and a circuit of signal processing. Normally, a rate transfer test and multi-position tests are us-ing for IMU calibration. The new calibration method is based on whole angle rotation or finite rotation. In fact it’s suggested to turn over IMU around three axes simultaneously. In order to solve the equation of calibration, it is necessary to provide an equality of a rank of basic matrix into degree of basic matrix. The results of simulated IMU data presented to demonstrate the performance of the new calibration method.

  1. Analysis and Automation of Calibration Process for Measurement Coils for Particle Accelerator Magnets

    CERN Document Server

    AUTHOR|(CDS)2226624; Azzam, Jamal; Chanthery, Elodie

    Various techniques are used to measure magnetic fields within the Magnetic Measurement (MM) section, all of which require the use of specific sensors. These sensors need to be calibrated in order to map the resulting signals to the characteristics of the magnetic field. Thus, the calibration procedure would benefit from being improved and made more stable, efficient, and technologically up-to-date. Consequently, to improve the efficiency of the calibration procedure and the work processes it was first necessary to analyze them and identify potential issues and weaknesses to be further addressed. The calibration procedure mainly suffered from too many manual operations, which were potential sources of error, and from outdated readout and data acquisition software and hardware. Firstly, a new software program had to be developed using a C++ framework called Flexible Framework for Magnetic Measurements (FFMM), that would automate the data acquisition and analysis. Next, a feasibility study had to be done ...

  2. User-friendly freehand ultrasound calibration using Lego bricks and automatic registration.

    Science.gov (United States)

    Xiao, Yiming; Yan, Charles Xiao Bo; Drouin, Simon; De Nigris, Dante; Kochanowska, Anna; Collins, D Louis

    2016-09-01

    As an inexpensive, noninvasive, and portable clinical imaging modality, ultrasound (US) has been widely employed in many interventional procedures for monitoring potential tissue deformation, surgical tool placement, and locating surgical targets. The application requires the spatial mapping between 2D US images and 3D coordinates of the patient. Although positions of the devices (i.e., ultrasound transducer) and the patient can be easily recorded by a motion tracking system, the spatial relationship between the US image and the tracker attached to the US transducer needs to be estimated through an US calibration procedure. Previously, various calibration techniques have been proposed, where a spatial transformation is computed to match the coordinates of corresponding features in a physical phantom and those seen in the US scans. However, most of these methods are difficult to use for novel users. We proposed an ultrasound calibration method by constructing a phantom from simple Lego bricks and applying an automated multi-slice 2D-3D registration scheme without volumetric reconstruction. The method was validated for its calibration accuracy and reproducibility. Our method yields a calibration accuracy of [Formula: see text] mm and a calibration reproducibility of 1.29 mm. We have proposed a robust, inexpensive, and easy-to-use ultrasound calibration method.

  3. An experimental procedure to determine heat transfer properties of turbochargers

    Science.gov (United States)

    Serrano, J. R.; Olmeda, P.; Páez, A.; Vidal, F.

    2010-03-01

    Heat transfer phenomena in turbochargers have been a subject of investigation due to their importance for the correct determination of compressor real work when modelling. The commonly stated condition of adiabaticity for turbochargers during normal operation of an engine has been revaluated because important deviations from adiabatic behaviour have been stated in many studies in this issue especially when the turbocharger is running at low rotational speeds/loads. The deviations mentioned do not permit us to assess properly the turbine and compressor efficiencies since the pure aerodynamic effects cannot be separated from the non-desired heat transfer due to the presence of both phenomena during turbocharger operation. The correction of the aforesaid facts is necessary to properly feed engine models with reliable information and in this way increase the quality of the results in any modelling process. The present work proposes a thermal characterization methodology successfully applied in a turbocharger for a passenger car which is based on the physics of the turbocharger. Its application helps to understand the thermal behaviour of the turbocharger, and the results obtained constitute vital information for future modelling efforts which involve the use of the information obtained from the proposed methodology. The conductance values obtained from the proposed methodology have been applied to correct a procedure for measuring the mechanical efficiency of the tested turbocharger.

  4. An experimental procedure to determine heat transfer properties of turbochargers

    International Nuclear Information System (INIS)

    Serrano, J R; Olmeda, P; Páez, A; Vidal, F

    2010-01-01

    Heat transfer phenomena in turbochargers have been a subject of investigation due to their importance for the correct determination of compressor real work when modelling. The commonly stated condition of adiabaticity for turbochargers during normal operation of an engine has been revaluated because important deviations from adiabatic behaviour have been stated in many studies in this issue especially when the turbocharger is running at low rotational speeds/loads. The deviations mentioned do not permit us to assess properly the turbine and compressor efficiencies since the pure aerodynamic effects cannot be separated from the non-desired heat transfer due to the presence of both phenomena during turbocharger operation. The correction of the aforesaid facts is necessary to properly feed engine models with reliable information and in this way increase the quality of the results in any modelling process. The present work proposes a thermal characterization methodology successfully applied in a turbocharger for a passenger car which is based on the physics of the turbocharger. Its application helps to understand the thermal behaviour of the turbocharger, and the results obtained constitute vital information for future modelling efforts which involve the use of the information obtained from the proposed methodology. The conductance values obtained from the proposed methodology have been applied to correct a procedure for measuring the mechanical efficiency of the tested turbocharger

  5. Pilot study to verify the calibration of electrometers

    International Nuclear Information System (INIS)

    Becker, P.; Meghzifene, A.

    2002-01-01

    Full text: The main detector used for standardization of the quantities used in measurements of ionizing radiation is the ionization chamber. The interaction of the radiation with this detector produces electrical charge, usually, in the range of pC to nC. The instrument used to measure such small charges (or currents) is the electrometer. As part of a good practice, the measured charge (current) must be traceable to a primary or secondary standard. Some calibration laboratories can only provide a system calibration coefficient, i.e. a calibration coefficient for the combination of electrometer plus ionization chamber (Gy/scale division). This practice is acceptable, but it can impose a limitation to the automation of their calibration procedures (using computerized application for the acquisition of current/charge). Not all models have the possibility of a connection to a computer and in the case of those that don't have this capability, automation is not possible without the development of a specific interface. In addition, end-users receive a calibration coefficient, which is only valid for the set ion chamber and electrometer. In case of a broken chamber, the end-user cannot connect another chamber to their electrometer without knowing its calibration coefficient. If the calibration laboratories had the capability of calibrating the chamber separately from the electrometer, for example, using an electrometer calibrated in terms of charge, all the chambers could be calibrated using this electrometer. The laboratory can also benefit from the automation of the measurements. This requires that the laboratory must be able to cross-calibrate the electrometers (associated to the chambers) also in terms of charge (Coulombs). Electrical charge is standardized by the use of a standard air capacitor and a standard voltage source (Q=CV) and the National Laboratory for Metrology of Ionizing Radiation (IRD) in Brazil has also adopted this procedure. Since the Brazilian

  6. LET calibration for CR-39 detectors in different oxygen environments

    International Nuclear Information System (INIS)

    Zhou, D.; Semones, E.; Weyland, M.; Benton, E.R.

    2007-01-01

    High LET (linear energy transfer) radiation is the main contributor to the radiation field in low Earth orbit (LEO) in terms of dose equivalent. CR-39 plastic nuclear track detectors (PNTDs) can measure the LET spectrum and charge spectrum for the complicated radiation field in space. Previous research indicated that the sensitivity of CR-39 is different for CR-39 PNTDs working in different oxygen environments. LET calibration for CR-39 detectors in different oxygen environments is needed. Almost all the previous LET calibration work was carried out for CR-39 detectors in good-oxygen condition, LET calibration work for CR-39 in poor-oxygen condition has not been conducted until our work. Systematic LET calibrations were carried out by JSC-SRAG (Space Radiation Analysis Group) for CR-39 detectors working in different oxygen environments and abundant results of LET calibrations were obtained. This paper introduces the method for CR-39 LET calibration, presents and discusses the calibration results and some applications

  7. 28 CFR 527.31 - Procedures.

    Science.gov (United States)

    2010-07-01

    ... 28 Judicial Administration 2 2010-07-01 2010-07-01 false Procedures. 527.31 Section 527.31 Judicial Administration BUREAU OF PRISONS, DEPARTMENT OF JUSTICE INMATE ADMISSION, CLASSIFICATION, AND TRANSFER TRANSFERS Transfer of Inmates to State Agents for Production on State Writs § 527.31 Procedures...

  8. CAMAC-controlled calibration system for nuclear reactor instruments

    International Nuclear Information System (INIS)

    McDowell, W.P.; Cornella, R.J.

    1977-01-01

    The hardware and the software which have been developed to implement a nuclear instrument calibration system for the Argonne National Laboratory ZPR-VI and ZPR-IX reactor complex are described. The system is implemented using an SEL-840 computer with its associated CAMAC crates and a hardware interface to generate input parameters and measure the required outputs on the instrument under test. Both linear and logarithmic instruments can be calibrated by the system and output parameters can be measured at various automatically selected values of ac line voltage. A complete report on each instrument is printed as a result of the calibration and out-of-tolerance readings are flagged. Operator interface is provided by a CAMAC-controlled Hazeltine terminal. The terminal display leads the operator through the complete calibration procedure. This computer-controlled system is a significant improvement over previously used methods of calibrating nuclear instruments since it reduces reactor downtime and allows rapid detection of long-term changes in instrument calibration

  9. Technical Note: Calibration device for the krypton hygrometer KH20

    Directory of Open Access Journals (Sweden)

    T. Foken

    2012-08-01

    Full Text Available A calibration device for krypton hygrometers (KH20, Campbell Scientific, Inc. with variable path length is presented. This unit allows for in-situ calibrations of the krypton hygrometer, which is typically not very stable over time, during measuring campaigns. It was constructed mainly for application at high altitudes and low temperatures, where further improvements are needed to the IR-hygrometers which are normally used. The changing path length requires that a changing concentration of the absorber be simulated. Because oxygen absorbs more strongly than water vapour, the calibration is made against oxygen and transferred to water vapour. The design of the calibration instrument is made as one unit containing a stepper motor system, PC and humidity sensor. For the calibration, it is necessary to install the krypton hygrometer on this unit.

  10. Calibration of optical cantilever deflection readers

    International Nuclear Information System (INIS)

    Hu Zhiyu; Seeley, Tim; Kossek, Sebastian; Thundat, Thomas

    2004-01-01

    Because of its ultrahigh sensitivity, the optical lever detection method similar to that used in the atomic force microscope (AFM) has been widely employed as a standard technique for measuring microcantilever deflection. Along with the increasing interest in using the microcantilever as a sensing platform, there is also a requirement for a reliable calibration technique. Many researchers have used the concept of optical lever detection to construct microcantilever deflection readout instruments for chemical, physical, and biological detection. However, without an AFM piezo z scanner, it is very difficult to precisely calibrate these instruments. Here, we present a step-by-step method to conveniently calibrate an instrument using commercially available piezoresistive cantilevers. The experimental results closely match the theoretical calculation. Following this procedure, one can easily calibrate any optical cantilever deflection detection system with high reproducibility, precision, and reliability. A detailed discussion of the optical lever readout system design has been addressed in this article

  11. Applications of the Italian protocol for the calibration of brachytherapy sources

    International Nuclear Information System (INIS)

    Piermattei, A.; Azario, L.

    1997-01-01

    The Associazione Italiana di Fisica Biomedica (AIFB; Italian Association of Biomedical Physics) has adopted the Italian protocol for the calibration of brachytherapy sources. The AIFB protocol allows measurements of the reference air kerma rate, dK/dt r , within 1.7% (1σ). To measure dK/dt r the AIFB protocol has identified a direct and an indirect procedure. The direct procedure is based on the use of spherical or cylindrical ionization chambers as local reference dosimeters positioned along the transverse bisector axis of the source. Once the source is specified by a dK/dt r value, this can be used to calibrate a field instrument, such as a well-type ionization chamber, for further source calibrations by means of an indirect procedure. This paper reports the results obtained by the Physics Laboratory of the Universita Cattolica del S Cuore (PL-UCSC), in terms of dK/dt r calibration of five types of source, 169 Yb, 192 Ir and 137 Cs. The role of the dK/dt r determination for a brachytherapy source has been underlined when a new source such as the 169 Yb seed model X1267 has been proposed for clinical use. The dK/dt r values for 137 Cs spherical sources differed by 5% from the vendor's mean value. The five types of source calibrated in terms of dK/dt r were used to obtain the calibration factor, N K r source , of an HDR-1000 well-type ionization chamber. (author)

  12. Calibration of photon and beta ray sources used in brachytherapy. Guidelines on standardized procedures at Secondary Standards Dosimetry Laboratories (SSDLs) and hospitals

    International Nuclear Information System (INIS)

    2002-03-01

    It has generally been recognized that international harmonization in radiotherapy dosimetry is essential. Consequently, the IAEA has given much effort to this, for example by publishing a number of reports in the Technical Reports Series (TRS) for external beam dosimetry, most notably TRS-277 and more recently TRS-398. Both of these reports describe in detail the steps to be taken for absorbed dose determination in water and they are often referred to as 'dosimetry protocols'. Similar to TRS-277, it is expected that TRS-398 will be adopted or used as a model by a large number of countries as their national protocol. In 1996, the IAEA established a calibration service for low dose rate (LDR) 137 Cs brachytherapy sources, which is the most widely used source for treatment of gynecological cancer. To further enhance harmonization in brachytherapy dosimetry, the IAEA published in 1999 IAEA-TECDOC-1079 entitled 'Calibration of Brachytherapy Sources. Guidelines on Standardized Procedures for the Calibration of Brachytherapy Sources at Secondary Standard Dosimetry Laboratories (SSDLs) and Hospitals'. The report was well received and was distributed in a large number of copies to the members of the IAEA/WHO network of SSDLs and to medical physicists working with brachytherapy. The present report is an update of the aforementioned TECDOC. Whereas TECDOC-1079 described methods for calibrating brachytherapy sources with photon energies at or above those of 192 Ir, the current report has a wider scope in that it deals with standardization of calibration of all the most commonly used brachytherapy sources, including both photon and beta emitting sources. The latter sources have been in use for a few decades already, but their calibration methods have been unclear. Methods are also described for calibrating sources used in the rapidly growing field of cardiovascular angioplasty. In this application, irradiation of the vessel wall is done in an attempt to prevent restenosis after

  13. Automatic Calibration of High Density Electric Muscle Stimulation

    DEFF Research Database (Denmark)

    Knibbe, Jarrod; Strohmeier, Paul; Boring, Sebastian

    2017-01-01

    . (2) EMS requires time consuming, expert calibration -- confining these interaction techniques to the lab. EMS arrays have been shown to increase stimulation resolution, but as calibration complexity increases exponentially as more electrodes are used, we require heuristics or automated procedures......Electric muscle stimulation (EMS) can enable mobile force feedback, support pedestrian navigation, or confer object affordances. To date, however, EMS is limited by two interlinked problems. (1) EMS is low resolution -- achieving only coarse movements and constraining opportunities for exploration...... for successful calibration. We explore the feasibility of using electromyography (EMG) to auto-calibrate high density EMS arrays. We determine regions of muscle activity during human-performed gestures, to inform stimulation patterns for EMS-performed gestures. We report on a study which shows that auto...

  14. Monitoring coordinate measuring machines by calibrated parts

    International Nuclear Information System (INIS)

    Weckenmann, A; Lorz, J

    2005-01-01

    Coordinate measuring machines (CMM) are essential for quality assurance and production control in modern manufacturing. Due to the necessity of assuring traceability during the use of CMM, interim checks with calibrated objects carried out periodically. For this purpose usually special artefacts like standardized ball plates, hole plates, ball bars or step gages are measured. Measuring calibrated series parts would be more advantageous. Applying the substitution method of ISO 15530-3: 2000 such parts can be used. It is less cost intensive and less time consuming than measuring expensive special standardized objects in special programmed measurement routines. Moreover, the measurement results can directly compare with the calibration values; thus, direct information on systematic measurement deviations and uncertainty of the measured features are available. The paper describes a procedure for monitoring horizontal-arm CMMs with calibrated sheet metal series parts

  15. Uranium cross-calibration measurements using an active well coincidence counter

    International Nuclear Information System (INIS)

    Nikolaev, V.; Prochine, I.; Smirnov, V.; Ensslin, N.; Carillo, L.

    1998-01-01

    This paper reports on the cross-calibration of an Active Well Coincidence Counter for use in the Materials Protection, Control, and Accountability Graduate Program at the Moscow State Engineering Physics Institute (MEPhI). The cross-calibration procedure and its application to nuclear material types available at MEPhI for instructional purposes is described. Cross-calibration results at Los Alamos and initial applications at MEPhI are summarized. Based on the results so far, the authors conclude that the cross-calibration approach seems useful, with good prospects for potential applications at other Russian and US Dept. of Energy facilities

  16. Automated magnification calibration in transmission electron microscopy using Fourier analysis of replica images

    International Nuclear Information System (INIS)

    Laak, Jeroen A.W.M. van der; Dijkman, Henry B.P.M.; Pahlplatz, Martin M.M.

    2006-01-01

    The magnification factor in transmission electron microscopy is not very precise, hampering for instance quantitative analysis of specimens. Calibration of the magnification is usually performed interactively using replica specimens, containing line or grating patterns with known spacing. In the present study, a procedure is described for automated magnification calibration using digital images of a line replica. This procedure is based on analysis of the power spectrum of Fourier transformed replica images, and is compared to interactive measurement in the same images. Images were used with magnification ranging from 1,000x to 200,000x. The automated procedure deviated on average 0.10% from interactive measurements. Especially for catalase replicas, the coefficient of variation of automated measurement was considerably smaller (average 0.28%) compared to that of interactive measurement (average 3.5%). In conclusion, calibration of the magnification in digital images from transmission electron microscopy may be performed automatically, using the procedure presented here, with high precision and accuracy

  17. Calibration Plans for the Global Precipitation Measurement (GPM)

    Science.gov (United States)

    Bidwell, S. W.; Flaming, G. M.; Adams, W. J.; Everett, D. F.; Mendelsohn, C. R.; Smith, E. A.; Turk, J.

    2002-01-01

    The Global Precipitation Measurement (GPM) is an international effort led by the National Aeronautics and Space Administration (NASA) of the U.S.A. and the National Space Development Agency of Japan (NASDA) for the purpose of improving research into the global water and energy cycle. GPM will improve climate, weather, and hydrological forecasts through more frequent and more accurate measurement of precipitation world-wide. Comprised of U.S. domestic and international partners, GPM will incorporate and assimilate data streams from many spacecraft with varied orbital characteristics and instrument capabilities. Two of the satellites will be provided directly by GPM, the core satellite and a constellation member. The core satellite, at the heart of GPM, is scheduled for launch in November 2007. The core will carry a conical scanning microwave radiometer, the GPM Microwave Imager (GMI), and a two-frequency cross-track-scanning radar, the Dual-frequency Precipitation Radar (DPR). The passive microwave channels and the two radar frequencies of the core are carefully chosen for investigating the varying character of precipitation over ocean and land, and from the tropics to the high-latitudes. The DPR will enable microphysical characterization and three-dimensional profiling of precipitation. The GPM-provided constellation spacecraft will carry a GMI radiometer identical to that on the core spacecraft. This paper presents calibration plans for the GPM, including on-board instrument calibration, external calibration methods, and the role of ground validation. Particular emphasis is on plans for inter-satellite calibration of the GPM constellation. With its Unique instrument capabilities, the core spacecraft will serve as a calibration transfer standard to the GPM constellation. In particular the Dual-frequency Precipitation Radar aboard the core will check the accuracy of retrievals from the GMI radiometer and will enable improvement of the radiometer retrievals

  18. The dissemination in France of the units used in ionizing radiation metrology

    International Nuclear Information System (INIS)

    Guiho, J.P.; Simoen, J.P.

    1978-01-01

    After reviewing the system of metrology in France the authors describe the working of the ionizing radiation calibration chain. Emphasis is laid on the procedures used for the transfer of the units of exposure and absorbed dose. Such transfers are carried out either by direct comparison with a standard kept at the calibration centre, or by special procedures involving the use of transfer dose meters or of radioactive sources calibrated and supplied by the primary laboratory or the calibration centre. An analysis of the steps and of the accumulation of errors is presented for each dosimetric quantity considered. The authors make a preliminary assessment of the operation of the French ionizing radiation calibration chain. (author)

  19. Calibration factor or calibration coefficient?

    International Nuclear Information System (INIS)

    Meghzifene, A.; Shortt, K.R.

    2002-01-01

    Full text: The IAEA/WHO network of SSDLs was set up in order to establish links between SSDL members and the international measurement system. At the end of 2001, there were 73 network members in 63 Member States. The SSDL network members provide calibration services to end-users at the national or regional level. The results of the calibrations are summarized in a document called calibration report or calibration certificate. The IAEA has been using the term calibration certificate and will continue using the same terminology. The most important information in a calibration certificate is a list of calibration factors and their related uncertainties that apply to the calibrated instrument for the well-defined irradiation and ambient conditions. The IAEA has recently decided to change the term calibration factor to calibration coefficient, to be fully in line with ISO [ISO 31-0], which recommends the use of the term coefficient when it links two quantities A and B (equation 1) that have different dimensions. The term factor should only be used for k when it is used to link the terms A and B that have the same dimensions A=k.B. However, in a typical calibration, an ion chamber is calibrated in terms of a physical quantity such as air kerma, dose to water, ambient dose equivalent, etc. If the chamber is calibrated together with its electrometer, then the calibration refers to the physical quantity to be measured per electrometer unit reading. In this case, the terms referred have different dimensions. The adoption by the Agency of the term coefficient to express the results of calibrations is consistent with the 'International vocabulary of basic and general terms in metrology' prepared jointly by the BIPM, IEC, ISO, OIML and other organizations. The BIPM has changed from factor to coefficient. The authors believe that this is more than just a matter of semantics and recommend that the SSDL network members adopt this change in terminology. (author)

  20. The calibration procedure of the radiation monitoring system installed in radiation controlled area of KOMAC

    Energy Technology Data Exchange (ETDEWEB)

    Park, Sung-Kyun; Min, Yi-Sub; Park, Jeong-Min; Cho, Yong-Sub [Korea Atomic Energy Research Institute, Gyeongju (Korea, Republic of)

    2016-10-15

    The spaces, where these accelerators are installed, are defined as the radiation controlled area and the levels of the radiation in this area are monitored by the radiation monitoring system (RMS) to protect radiation workers and experiment users from the hazards of the ionizing radiation and the surface and air contamination tests are carried out periodically by the radiation secure team. The most of RMS instruments are installed in the accelerator building, where the 100-MeV proton linear accelerator is installed. All detectors of RMS should be calibrated every year to prove the reliability of RMS and almost all instruments for RMS was calibrated during this summer maintenance period of KOMAC this year. Almost all RMS instruments installed in KOMAC is calibrated between 2016-07-13 and 2016-08-24. As the calibration result, if the current reading value are within the 5% of the reference dose rate value, this RMS instrument can be used one more year. Otherwise, the detector of that RMS instrument should be repaired or replaced. The self-calibration certificate for each RMS instrument will be published only for the instrument to satisfy the condition.

  1. SCALA: In situ calibration for integral field spectrographs

    Science.gov (United States)

    Lombardo, S.; Küsters, D.; Kowalski, M.; Aldering, G.; Antilogus, P.; Bailey, S.; Baltay, C.; Barbary, K.; Baugh, D.; Bongard, S.; Boone, K.; Buton, C.; Chen, J.; Chotard, N.; Copin, Y.; Dixon, S.; Fagrelius, P.; Feindt, U.; Fouchez, D.; Gangler, E.; Hayden, B.; Hillebrandt, W.; Hoffmann, A.; Kim, A. G.; Leget, P.-F.; McKay, L.; Nordin, J.; Pain, R.; Pécontal, E.; Pereira, R.; Perlmutter, S.; Rabinowitz, D.; Reif, K.; Rigault, M.; Rubin, D.; Runge, K.; Saunders, C.; Smadja, G.; Suzuki, N.; Taubenberger, S.; Tao, C.; Thomas, R. C.; Nearby Supernova Factory

    2017-11-01

    Aims: The scientific yield of current and future optical surveys is increasingly limited by systematic uncertainties in the flux calibration. This is the case for type Ia supernova (SN Ia) cosmology programs, where an improved calibration directly translates into improved cosmological constraints. Current methodology rests on models of stars. Here we aim to obtain flux calibration that is traceable to state-of-the-art detector-based calibration. Methods: We present the SNIFS Calibration Apparatus (SCALA), a color (relative) flux calibration system developed for the SuperNova integral field spectrograph (SNIFS), operating at the University of Hawaii 2.2 m (UH 88) telescope. Results: By comparing the color trend of the illumination generated by SCALA during two commissioning runs, and to previous laboratory measurements, we show that we can determine the light emitted by SCALA with a long-term repeatability better than 1%. We describe the calibration procedure necessary to control for system aging. We present measurements of the SNIFS throughput as estimated by SCALA observations. Conclusions: The SCALA calibration unit is now fully deployed at the UH 88 telescope, and with it color-calibration between 4000 Å and 9000 Å is stable at the percent level over a one-year baseline.

  2. Calibration procedures for the quantitative determination of membrane potential in human cells using anionic dyes.

    Science.gov (United States)

    Klapperstück, Thomas; Glanz, Dagobert; Hanitsch, Stefan; Klapperstück, Manuela; Markwardt, Fritz; Wohlrab, Johannes

    2013-07-01

    Quantitative determinations of the cell membrane potential of lymphocytes (Wilson et al., J Cell Physiol 1985;125:72-81) and thymocytes (Krasznai et al., J Photochem Photobiol B 1995;28:93-99) using the anionic dye DiBAC4 (3) proved that dye depletion in the extracellular medium as a result of cellular uptake can be negligible over a wide range of cell densities. In contrast, most flow cytometric studies have not verified this condition but rather assumed it from the start. Consequently, the initially prepared extracellular dye concentration has usually been used for the calculation of the Nernst potential of the dye. In this study, however, external dye depletion could be observed in both large IGR-1 and small LCL-HO cells under experimental conditions, which have often been applied routinely in spectrofluorimetry and flow cytometry. The maximum cell density at which dye depletion could be virtually avoided was dependent on cell size and membrane potential and definitely needed to be taken into account to ensure reliable results. In addition, accepted calibration procedures based on the partition of sodium and potassium (Goldman-Hodgkin-Katz equation) or potassium alone (Nernst equation) were performed by flow cytometry on cell suspensions with an appropriately low cell density. The observed extensive lack of concordance between the correspondingly calculated membrane potential and the equilibrium potential of DiBAC4 (3) revealed that these methods require the additional measurement of cation parameters (membrane permeability and/or intracellular concentration). In contrast, due to the linear relation between fluorescence and low DiBAC4 (3) concentrations, the Nernst potential of the dye for totally depolarized cells can be reliably used for calibration with an essentially lower effort and expense. Copyright © 2013 International Society for Advancement of Cytometry.

  3. Calibration of Ge gamma-ray spectrometers for complex sample geometries and matrices

    Energy Technology Data Exchange (ETDEWEB)

    Semkow, T.M., E-mail: thomas.semkow@health.ny.gov [Wadsworth Center, New York State Department of Health, Empire State Plaza, Albany, NY 12201 (United States); Department of Environmental Health Sciences, School of Public Health, University at Albany, State University of New York, Rensselaer, NY 12144 (United States); Bradt, C.J.; Beach, S.E.; Haines, D.K.; Khan, A.J.; Bari, A.; Torres, M.A.; Marrantino, J.C.; Syed, U.-F. [Wadsworth Center, New York State Department of Health, Empire State Plaza, Albany, NY 12201 (United States); Kitto, M.E. [Wadsworth Center, New York State Department of Health, Empire State Plaza, Albany, NY 12201 (United States); Department of Environmental Health Sciences, School of Public Health, University at Albany, State University of New York, Rensselaer, NY 12144 (United States); Hoffman, T.J. [Wadsworth Center, New York State Department of Health, Empire State Plaza, Albany, NY 12201 (United States); Curtis, P. [Kiltel Systems, Inc., Clyde Hill, WA 98004 (United States)

    2015-11-01

    A comprehensive study of the efficiency calibration and calibration verification of Ge gamma-ray spectrometers was performed using semi-empirical, computational Monte-Carlo (MC), and transfer methods. The aim of this study was to evaluate the accuracy of the quantification of gamma-emitting radionuclides in complex matrices normally encountered in environmental and food samples. A wide range of gamma energies from 59.5 to 1836.0 keV and geometries from a 10-mL jar to 1.4-L Marinelli beaker were studied on four Ge spectrometers with the relative efficiencies between 102% and 140%. Density and coincidence summing corrections were applied. Innovative techniques were developed for the preparation of artificial complex matrices from materials such as acidified water, polystyrene, ethanol, sugar, and sand, resulting in the densities ranging from 0.3655 to 2.164 g cm{sup −3}. They were spiked with gamma activity traceable to international standards and used for calibration verifications. A quantitative method of tuning MC calculations to experiment was developed based on a multidimensional chi-square paraboloid. - Highlights: • Preparation and spiking of traceable complex matrices in extended geometries. • Calibration of Ge gamma spectrometers for complex matrices. • Verification of gamma calibrations. • Comparison of semi-empirical, computational Monte Carlo, and transfer methods of Ge calibration. • Tuning of Monte Carlo calculations using a multidimensional paraboloid.

  4. Calibration of Ge gamma-ray spectrometers for complex sample geometries and matrices

    International Nuclear Information System (INIS)

    Semkow, T.M.; Bradt, C.J.; Beach, S.E.; Haines, D.K.; Khan, A.J.; Bari, A.; Torres, M.A.; Marrantino, J.C.; Syed, U.-F.; Kitto, M.E.; Hoffman, T.J.; Curtis, P.

    2015-01-01

    A comprehensive study of the efficiency calibration and calibration verification of Ge gamma-ray spectrometers was performed using semi-empirical, computational Monte-Carlo (MC), and transfer methods. The aim of this study was to evaluate the accuracy of the quantification of gamma-emitting radionuclides in complex matrices normally encountered in environmental and food samples. A wide range of gamma energies from 59.5 to 1836.0 keV and geometries from a 10-mL jar to 1.4-L Marinelli beaker were studied on four Ge spectrometers with the relative efficiencies between 102% and 140%. Density and coincidence summing corrections were applied. Innovative techniques were developed for the preparation of artificial complex matrices from materials such as acidified water, polystyrene, ethanol, sugar, and sand, resulting in the densities ranging from 0.3655 to 2.164 g cm −3 . They were spiked with gamma activity traceable to international standards and used for calibration verifications. A quantitative method of tuning MC calculations to experiment was developed based on a multidimensional chi-square paraboloid. - Highlights: • Preparation and spiking of traceable complex matrices in extended geometries. • Calibration of Ge gamma spectrometers for complex matrices. • Verification of gamma calibrations. • Comparison of semi-empirical, computational Monte Carlo, and transfer methods of Ge calibration. • Tuning of Monte Carlo calculations using a multidimensional paraboloid

  5. Temperature calibration procedure for thin film substrates for thermo-ellipsometric analysis using melting point standards

    NARCIS (Netherlands)

    Kappert, Emiel; Raaijmakers, Michiel; Ogieglo, Wojciech; Nijmeijer, Arian; Huiskes, Cindy; Huiskes, C.; Benes, Nieck Edwin

    2015-01-01

    Precise and accurate temperature control is pertinent to studying thermally activated processes in thin films. Here, we present a calibration method for the substrate–film interface temperature using spectroscopic ellipsometry. The method is adapted from temperature calibration methods that are well

  6. CALIBRATION OF LOW COST DIGITAL CAMERA USING DATA FROM SIMULTANEOUS LIDAR AND PHOTOGRAMMETRIC SURVEYS

    Directory of Open Access Journals (Sweden)

    E. Mitishita

    2012-07-01

    Full Text Available Digital photogrammetric products from the integration of imagery and lidar datasets are a reality nowadays. When the imagery and lidar surveys are performed together and the camera is connected to the lidar system, a direct georeferencing can be applied to compute the exterior orientation parameters of the images. Direct georeferencing of the images requires accurate interior orientation parameters to perform photogrammetric application. Camera calibration is a procedure applied to compute the interior orientation parameters (IOPs. Calibration researches have established that to obtain accurate IOPs, the calibration must be performed with same or equal condition that the photogrammetric survey is done. This paper shows the methodology and experiments results from in situ self-calibration using a simultaneous images block and lidar dataset. The calibration results are analyzed and discussed. To perform this research a test field was fixed in an urban area. A set of signalized points was implanted on the test field to use as the check points or control points. The photogrammetric images and lidar dataset of the test field were taken simultaneously. Four strips of flight were used to obtain a cross layout. The strips were taken with opposite directions of flight (W-E, E-W, N-S and S-N. The Kodak DSC Pro SLR/c digital camera was connected to the lidar system. The coordinates of the exposition station were computed from the lidar trajectory. Different layouts of vertical control points were used in the calibration experiments. The experiments use vertical coordinates from precise differential GPS survey or computed by an interpolation procedure using the lidar dataset. The positions of the exposition stations are used as control points in the calibration procedure to eliminate the linear dependency of the group of interior and exterior orientation parameters. This linear dependency happens, in the calibration procedure, when the vertical images and

  7. Tanks for liquids: calibration and errors assessment

    International Nuclear Information System (INIS)

    Espejo, J.M.; Gutierrez Fernandez, J.; Ortiz, J.

    1980-01-01

    After a brief reference to some of the problems raised by tanks calibration, two methods, theoretical and experimental are presented, so as to achieve it taking into account measurement errors. The method is applied to the transfer of liquid from one tank to another. Further, a practical example is developed. (author)

  8. Measurement quality assurance for beta particle calibrations at NIST

    Energy Technology Data Exchange (ETDEWEB)

    Soares, C.G.; Pruitt, J.S. [National Institute of Standards and Technology, Gaithersburg, MD (United States)

    1993-12-31

    Standardized beta-particle fields have been established in an international standard and have been adopted for use in several U.S. dosimeter and instrument testing standards. Calibration methods and measurement quality assurance procedures employed at the National Institute of Standards and Technology (NIST) for beta-particle calibrations in these reference fields are discussed. The calibration facility including the NIST-automated extrapolation ionization chamber is described, and some sample results of calibrations are shown. Methods for establishing and maintaining traceability to NIST of secondary laboratories are discussed. Currently, there are problems in finding a good method for routine testing of traceability to NIST. Some examples of past testing methods are given and solutions to this problem are proposed.

  9. Measurement quality assurance for beta particle calibrations at NIST

    International Nuclear Information System (INIS)

    Soares, C.G.; Pruitt, J.S.

    1993-01-01

    Standardized beta-particle fields have been established in an international standard and have been adopted for use in several U.S. dosimeter and instrument testing standards. Calibration methods and measurement quality assurance procedures employed at the National Institute of Standards and Technology (NIST) for beta-particle calibrations in these reference fields are discussed. The calibration facility including the NIST-automated extrapolation ionization chamber is described, and some sample results of calibrations are shown. Methods for establishing and maintaining traceability to NIST of secondary laboratories are discussed. Currently, there are problems in finding a good method for routine testing of traceability to NIST. Some examples of past testing methods are given and solutions to this problem are proposed

  10. One step geometrical calibration method for optical coherence tomography

    International Nuclear Information System (INIS)

    Díaz, Jesús Díaz; Ortmaier, Tobias; Stritzel, Jenny; Rahlves, Maik; Reithmeier, Eduard; Roth, Bernhard; Majdani, Omid

    2016-01-01

    We present a novel one-step calibration methodology for geometrical distortion correction for optical coherence tomography (OCT). A calibration standard especially designed for OCT is introduced, which consists of an array of inverse pyramidal structures. The use of multiple landmarks situated on four different height levels on the pyramids allow performing a 3D geometrical calibration. The calibration procedure itself is based on a parametric model of the OCT beam propagation. It is validated by experimental results and enables the reduction of systematic errors by more than one order of magnitude. In future, our results can improve OCT image reconstruction and interpretation for medical applications such as real time monitoring of surgery. (paper)

  11. Cs-137 brachytherapy sources calibration with well chamber

    International Nuclear Information System (INIS)

    Brunetto, M.; Sansogne, R.; Arbiser, S.; Duran, M.P.

    2004-01-01

    This work describes the procedures and actions developed for the identification and reference air kerma rate (S k ) verification of Cs-137 sources used in gynecological brachytherapy practices. Following the IAEA TECDOC 1151 recommendations, the first stage consisted in designing the documentation required for the inventory and shipping registry of sources, along with the digital spreadsheets for calculating the decay and S k of the sources at the moment of implantation. As a second stage, the S k of sources was measured, following the low dose rate sources protocol advise, with a Standard Imaging HDR 1000 Plus well chamber calibrated at the University of Wisconsin SSDL. The documentation generated through this procedure allows identify each source clearly and uni-vocally. No significant differences were found between the S k values obtained from the well chamber calibration procedure and those reported by the manufacturer in the corresponding certificates. The highest percent difference found was 2.3%. (author) [es

  12. Calibration of hydrological model with programme PEST

    Science.gov (United States)

    Brilly, Mitja; Vidmar, Andrej; Kryžanowski, Andrej; Bezak, Nejc; Šraj, Mojca

    2016-04-01

    PEST is tool based on minimization of an objective function related to the root mean square error between the model output and the measurement. We use "singular value decomposition", section of the PEST control file, and Tikhonov regularization method for successfully estimation of model parameters. The PEST sometimes failed if inverse problems were ill-posed, but (SVD) ensures that PEST maintains numerical stability. The choice of the initial guess for the initial parameter values is an important issue in the PEST and need expert knowledge. The flexible nature of the PEST software and its ability to be applied to whole catchments at once give results of calibration performed extremely well across high number of sub catchments. Use of parallel computing version of PEST called BeoPEST was successfully useful to speed up calibration process. BeoPEST employs smart slaves and point-to-point communications to transfer data between the master and slaves computers. The HBV-light model is a simple multi-tank-type model for simulating precipitation-runoff. It is conceptual balance model of catchment hydrology which simulates discharge using rainfall, temperature and estimates of potential evaporation. Version of HBV-light-CLI allows the user to run HBV-light from the command line. Input and results files are in XML form. This allows to easily connecting it with other applications such as pre and post-processing utilities and PEST itself. The procedure was applied on hydrological model of Savinja catchment (1852 km2) and consists of twenty one sub-catchments. Data are temporary processed on hourly basis.

  13. Monitoring scanner calibration using the image-derived arterial blood SUV in whole-body FDG-PET.

    Science.gov (United States)

    Maus, Jens; Hofheinz, Frank; Apostolova, Ivayla; Kreissl, Michael C; Kotzerke, Jörg; van den Hoff, Jörg

    2018-05-15

    The current de facto standard for quantification of tumor metabolism in oncological whole-body PET is the standardized uptake value (SUV) approach. SUV determination requires accurate scanner calibration. Residual inaccuracies of the calibration lead to biased SUV values. Especially, this can adversely affect multicenter trials where it is difficult to ensure reliable cross-calibration across participating sites. The goal of the present work was the evaluation of a new method for monitoring scanner calibration utilizing the image-derived arterial blood SUV (BSUV) averaged over a sufficiently large number of whole-body FDG-PET investigations. Data of 681 patients from three sites which underwent routine 18 F-FDG PET/CT or PET/MR were retrospectively analyzed. BSUV was determined in the descending aorta using a three-dimensional ROI concentric to the aorta's centerline. The ROI was delineated in the CT or MRI images and transferred to the PET images. A minimum ROI volume of 5 mL and a concentric safety margin to the aortic wall was observed. Mean BSUV, standard deviation (SD), and standard error of the mean (SE) were computed for three groups of patients at each site, investigated 2 years apart, respectively, with group sizes between 53 and 100 patients. Differences of mean BSUV between the individual groups and sites were determined. SD (SE) of BSUV in the different groups ranged from 14.3 to 20.7% (1.7 to 2.8%). Differences of mean BSUV between intra-site groups were small (1.1-6.3%). Only one out of nine of these differences reached statistical significance. Inter-site differences were distinctly larger (12.6-25.1%) and highly significant (PPET investigations is a viable approach for ensuring consistent scanner calibration over time and across different sites. We propose this approach as a quality control and cross-calibration tool augmenting established phantom-based procedures.

  14. E143 experiment. Shower counter calibration

    International Nuclear Information System (INIS)

    Fonvieille, H.; Grenier, P.

    1994-01-01

    The calibration procedure for the shower counters used in the E143 experiment is described. It has been developed during january 1994 in view of being used for the quick analysis. The method is explained and the results obtained on a given run are presented. (author)

  15. Calibration and Validation of the National Ecological Observatory Network's Airborne Imaging Spectrometers

    Science.gov (United States)

    Leisso, N.

    2015-12-01

    The National Ecological Observatory Network (NEON) is being constructed by the National Science Foundation and is slated for completion in 2017. NEON is designed to collect data to improve the understanding of changes in observed ecosystems. The observatory will produce data products on a variety of spatial and temporal scales collected from individual sites strategically located across the U.S. including Alaska, Hawaii, and Puerto Rico. Data sources include standardized terrestrial, instrumental, and aquatic observation systems in addition to three airborne remote sensing observation systems installed into leased Twin Otter aircraft. The Airborne Observation Platforms (AOP) are designed to collect 3-band aerial imagery, waveform and discrete LiDAR, and high-fidelity imaging spectroscopy data over the NEON sites annually at or near peak-greenness. The NEON Imaging Spectrometer (NIS) is a Visible and Shortwave Infrared (VSWIR) sensor designed by NASA JPL for ecological applications. Spectroscopic data is collected at 5-nm intervals across the solar-reflective spectral region (380-nm to 2500-nm) in a 34-degree FOV swath. A key uncertainty driver to the derived remote sensing NEON data products is the calibration of the imaging spectrometers. In addition, the calibration and accuracy of the higher-level data product algorithms is essential to the overall NEON mission to detect changes in the collected ecosystems over the 30-year expected lifetime. The typical calibration workflow of the NIS consists of the characterizing the focal plane, spectral calibration, and radiometric calibration. Laboratory spectral calibration is based on well-defined emission lines in conjunction with a scanning monochromator to define the individual spectral response functions. The radiometric calibration is NIST traceable and transferred to the NIS with an integrating sphere calibrated through the use of transfer radiometers. The laboratory calibration is monitored and maintained through

  16. Color calibration and color-managed medical displays: does the calibration method matter?

    Science.gov (United States)

    Roehrig, Hans; Rehm, Kelly; Silverstein, Louis D.; Dallas, William J.; Fan, Jiahua; Krupinski, Elizabeth A.

    2010-02-01

    Our laboratory has investigated the efficacy of a suite of color calibration and monitor profiling packages which employ a variety of color measurement sensors. Each of the methods computes gamma correction tables for the red, green and blue color channels of a monitor that attempt to: a) match a desired luminance range and tone reproduction curve; and b) maintain a target neutral point across the range of grey values. All of the methods examined here produce International Color Consortium (ICC) profiles that describe the color rendering capabilities of the monitor after calibration. Color profiles incorporate a transfer matrix that establishes the relationship between RGB driving levels and the International Commission on Illumination (CIE) XYZ (tristimulus) values of the resulting on-screen color; the matrix is developed by displaying color patches of known RGB values on the monitor and measuring the tristimulus values with a sensor. The number and chromatic distribution of color patches varies across methods and is usually not under user control. In this work we examine the effect of employing differing calibration and profiling methods on rendition of color images. A series of color patches encoded in sRGB color space were presented on the monitor using color-management software that utilized the ICC profile produced by each method. The patches were displayed on the calibrated monitor and measured with a Minolta CS200 colorimeter. Differences in intended and achieved luminance and chromaticity were computed using the CIE DE2000 color-difference metric, in which a value of ▵E = 1 is generally considered to be approximately one just noticeable difference (JND) in color. We observed between one and 17 JND's for individual colors, depending on calibration method and target.

  17. Third-party brachytherapy source calibrations and physicist responsibilities: Report of the AAPM Low Energy Brachytherapy Source Calibration Working Group

    International Nuclear Information System (INIS)

    Butler, Wayne M.; Bice, William S. Jr.; DeWerd, Larry A.; Hevezi, James M.; Huq, M. Saiful; Ibbott, Geoffrey S.; Palta, Jatinder R.; Rivard, Mark J.; Seuntjens, Jan P.; Thomadsen, Bruce R.

    2008-01-01

    The AAPM Low Energy Brachytherapy Source Calibration Working Group was formed to investigate and recommend quality control and quality assurance procedures for brachytherapy sources prior to clinical use. Compiling and clarifying recommendations established by previous AAPM Task Groups 40, 56, and 64 were among the working group's charges, which also included the role of third-party handlers to perform loading and assay of sources. This document presents the findings of the working group on the responsibilities of the institutional medical physicist and a clarification of the existing AAPM recommendations in the assay of brachytherapy sources. Responsibility for the performance and attestation of source assays rests with the institutional medical physicist, who must use calibration equipment appropriate for each source type used at the institution. Such equipment and calibration procedures shall ensure secondary traceability to a national standard. For each multi-source implant, 10% of the sources or ten sources, whichever is greater, are to be assayed. Procedures for presterilized source packaging are outlined. The mean source strength of the assayed sources must agree with the manufacturer's stated strength to within 3%, or action must be taken to resolve the difference. Third party assays do not absolve the institutional physicist from the responsibility to perform the institutional measurement and attest to the strength of the implanted sources. The AAPM leaves it to the discretion of the institutional medical physicist whether the manufacturer's or institutional physicist's measured value should be used in performing dosimetry calculations

  18. CryoSat-2 SIRAL Calibration and Performance

    Science.gov (United States)

    Fornari, M.; Scagliola, M.; Tagliani, N.; Parrinello, T.

    2012-12-01

    The main payload of CryoSat-2 is a Ku band pulse-width limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for azimuth processing. This allows to reach an along track resolution of about 250 meters which is a significant improvement over traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed, including both an internal and external calibration. The internal calibration monitors the instrument impulse response and the transfer function, like traditional altimeters. In addition to that, the interferometer requires a special calibration developed ad hoc for SIRAL. The external calibration is performed with the use of a ground transponder, located in Svalbard, which receives SIRAL signal and sends the echo back to the satellite. Internal calibration data are processed on ground by the CryoSat-2 Instrument Processing Facility (IPF1) and then applied to the science data. In December 2012, two and a half years of calibration data will be available, which will be shown in this poster. The external calibration (transponder) data are processed and analyzed independently from the operational chain. The use of an external transponder has been very useful to determine instrument performance and for the tuning of the on-ground processor. This poster presents the transponder results in terms of range noise and datation error.

  19. Employing an Incentive Spirometer to Calibrate Tidal Volumes Estimated from a Smartphone Camera

    Science.gov (United States)

    Reyes, Bersain A.; Reljin, Natasa; Kong, Youngsun; Nam, Yunyoung; Ha, Sangho; Chon, Ki H.

    2016-01-01

    A smartphone-based tidal volume (VT) estimator was recently introduced by our research group, where an Android application provides a chest movement signal whose peak-to-peak amplitude is highly correlated with reference VT measured by a spirometer. We found a Normalized Root Mean Squared Error (NRMSE) of 14.998% ± 5.171% (mean ± SD) when the smartphone measures were calibrated using spirometer data. However, the availability of a spirometer device for calibration is not realistic outside clinical or research environments. In order to be used by the general population on a daily basis, a simple calibration procedure not relying on specialized devices is required. In this study, we propose taking advantage of the linear correlation between smartphone measurements and VT to obtain a calibration model using information computed while the subject breathes through a commercially-available incentive spirometer (IS). Experiments were performed on twelve (N = 12) healthy subjects. In addition to corroborating findings from our previous study using a spirometer for calibration, we found that the calibration procedure using an IS resulted in a fixed bias of −0.051 L and a RMSE of 0.189 ± 0.074 L corresponding to 18.559% ± 6.579% when normalized. Although it has a small underestimation and slightly increased error, the proposed calibration procedure using an IS has the advantages of being simple, fast, and affordable. This study supports the feasibility of developing a portable smartphone-based breathing status monitor that provides information about breathing depth, in addition to the more commonly estimated respiratory rate, on a daily basis. PMID:26999152

  20. CP-50 calibration facility radiological safety assessment document

    International Nuclear Information System (INIS)

    Chilton, M.W.; Hill, R.L.; Eubank, B.F.

    1980-03-01

    The CP-50 Calibration Facility Radiological Safety Assessment document, prepared at the request of the Nevada Operations Office of the US Department of Energy to satisfy provisions of ERDA Manual Chapter 0531, presents design features, systems controls, and procedures used in the operation of the calibration facility. Site and facility characteristics and routine and non-routine operations, including hypothetical incidents or accidents are discussed and design factors, source control systems, and radiation monitoring considerations are described

  1. Transient, compressible heat and mass transfer in porous media using the strongly implicit iteration procedure.

    Science.gov (United States)

    Curry, D. M.; Cox, J. E.

    1972-01-01

    Coupled nonlinear partial differential equations describing heat and mass transfer in a porous matrix are solved in finite difference form with the aid of a new iterative technique (the strongly implicit procedure). Example numerical results demonstrate the characteristics of heat and mass transport in a porous matrix such as a charring ablator. It is emphasized that multidimensional flow must be considered when predicting the thermal response of a porous material subjected to nonuniform boundary conditions.

  2. Calibration of an electron volt neutron spectrometer

    International Nuclear Information System (INIS)

    Mayers, J.; Adams, M.A.

    2011-01-01

    The procedure for calibrating the VESUVIO eV neutron spectrometer at the ISIS neutron source is described. VESUVIO is used primarily to measure the momentum distribution n(p) of atoms, by inelastic scattering of very high energy (5-150 eV) neutrons. The results of the calibrations show that measurements of n(p) in atoms with masses lower than 16 amu can be measured with a resolution width ∼25% of the intrinsic peak widths in the current instrument configuration. Some suggestions as to how the instrument resolution could be significantly improved are made.

  3. Kinematic parameter calibration method for industrial robot manipulator using the relative position

    International Nuclear Information System (INIS)

    Ha, In Chul

    2008-01-01

    A new calibration method for industrial robot system calibration on a manufacturing floor is presented in this paper. To calibrate the robot system, a laser sensor to measure the distance between robot tool and measurement surface is attached to the robot end-effector and a grid is established in the floor. Given two position command pulses for a robot manipulator and using the position difference between two command pulses, the relative position measurement calibration method will find the real robot kinematic parameters. The procedures developed have been applied to an industrial robot. Finally, the effects of the models used to calibrate the robot are discussed. This calibration method represents an effective, low cost and feasible technique for the industrial robot calibration in lab. projects and industrial environments

  4. Calibration of laboratory equipment and its intermediate verification

    International Nuclear Information System (INIS)

    Remedi, Jorge O.

    2011-01-01

    When a laboratory wants to prove that he has technical competence to carry out tests or calibrations must demonstrate that it has complied with certain requirements that establish , among others, the mandatory : calibrate or verify equipment before putting it into service in order to ensure that it meets to the specifications of laboratory equipment to keep records evidencing the checks that equipment complies with the specification ; perform intermediate checks for maintain confidence in the calibration status of the equipment , ensure that the operation is checked and calibration status of equipment when the equipment goes outside the direct control of the laboratory , before be returned to service, establish a program and procedure for the calibration of equipment; show how determined the calibration periods of their equipment as well as evidence that intermediate checks are suitable for the calibration periods. However, some confusion is observed as to the meaning of the terms 'calibration' and 'verification' of a computer. This paper analyzes applicable documentation and suggests that the differences are generated in part by translations and by characterization concepts upon its usage, that is, if it is legal metrology or assessment conformity. Therefore, this study aims to characterize both concepts , fundamentals to zoom distinguish , outline appropriate strategies for calibration and verification activities to ensure the compliance with regulatory requirements [es

  5. A study of batch calibrations on 4 inch open faced charcoal adsorbers from four different manufacturers and how they compare to EERF's published calibration curves

    International Nuclear Information System (INIS)

    Jones, D.L.; Howell, T.E.

    1990-01-01

    The EERF Standard Operating Procedures for Rn-222 Measurements Using Charcoal Canisters is quite explicit in stating that the calibration tables and curves published in this manual are only typical calibration data that were developed for a particular batch of canisters from a supplier, and that new calibration data are developed and used with each different lot of canisters. In this paper, batch calibrations were performed in the authors' radon chambers on so called EPA style canisters from four different companies. The authors then prepared calibration tables and curves for each batch and compared them to the EERF published tables. it was our premise that EPA never intended for their curves to be considered standard and that with more and more companies manufacturing canisters, the calibration factors could certainly vary and might significantly differ from those derived by EERF several years ago

  6. Cross-correlation redshift calibration without spectroscopic calibration samples in DES Science Verification Data

    Science.gov (United States)

    Davis, C.; Rozo, E.; Roodman, A.; Alarcon, A.; Cawthon, R.; Gatti, M.; Lin, H.; Miquel, R.; Rykoff, E. S.; Troxel, M. A.; Vielzeuf, P.; Abbott, T. M. C.; Abdalla, F. B.; Allam, S.; Annis, J.; Bechtol, K.; Benoit-Lévy, A.; Bertin, E.; Brooks, D.; Buckley-Geer, E.; Burke, D. L.; Carnero Rosell, A.; Carrasco Kind, M.; Carretero, J.; Castander, F. J.; Crocce, M.; Cunha, C. E.; D'Andrea, C. B.; da Costa, L. N.; Desai, S.; Diehl, H. T.; Doel, P.; Drlica-Wagner, A.; Fausti Neto, A.; Flaugher, B.; Fosalba, P.; Frieman, J.; García-Bellido, J.; Gaztanaga, E.; Gerdes, D. W.; Giannantonio, T.; Gruen, D.; Gruendl, R. A.; Gutierrez, G.; Honscheid, K.; Jain, B.; James, D. J.; Jeltema, T.; Krause, E.; Kuehn, K.; Kuhlmann, S.; Kuropatkin, N.; Lahav, O.; Li, T. S.; Lima, M.; March, M.; Marshall, J. L.; Martini, P.; Melchior, P.; Ogando, R. L. C.; Plazas, A. A.; Romer, A. K.; Sanchez, E.; Scarpine, V.; Schindler, R.; Schubnell, M.; Sevilla-Noarbe, I.; Smith, M.; Soares-Santos, M.; Sobreira, F.; Suchyta, E.; Swanson, M. E. C.; Tarle, G.; Thomas, D.; Vikram, V.; Walker, A. R.; Wechsler, R. H.

    2018-06-01

    Galaxy cross-correlations with high-fidelity redshift samples hold the potential to precisely calibrate systematic photometric redshift uncertainties arising from the unavailability of complete and representative training and validation samples of galaxies. However, application of this technique in the Dark Energy Survey (DES) is hampered by the relatively low number density, small area, and modest redshift overlap between photometric and spectroscopic samples. We propose instead using photometric catalogues with reliable photometric redshifts for photo-z calibration via cross-correlations. We verify the viability of our proposal using redMaPPer clusters from the Sloan Digital Sky Survey (SDSS) to successfully recover the redshift distribution of SDSS spectroscopic galaxies. We demonstrate how to combine photo-z with cross-correlation data to calibrate photometric redshift biases while marginalizing over possible clustering bias evolution in either the calibration or unknown photometric samples. We apply our method to DES Science Verification (DES SV) data in order to constrain the photometric redshift distribution of a galaxy sample selected for weak lensing studies, constraining the mean of the tomographic redshift distributions to a statistical uncertainty of Δz ˜ ±0.01. We forecast that our proposal can, in principle, control photometric redshift uncertainties in DES weak lensing experiments at a level near the intrinsic statistical noise of the experiment over the range of redshifts where redMaPPer clusters are available. Our results provide strong motivation to launch a programme to fully characterize the systematic errors from bias evolution and photo-z shapes in our calibration procedure.

  7. Hydrological processes and model representation: impact of soft data on calibration

    Science.gov (United States)

    J.G. Arnold; M.A. Youssef; H. Yen; M.J. White; A.Y. Sheshukov; A.M. Sadeghi; D.N. Moriasi; J.L. Steiner; Devendra Amatya; R.W. Skaggs; E.B. Haney; J. Jeong; M. Arabi; P.H. Gowda

    2015-01-01

    Hydrologic and water quality models are increasingly used to determine the environmental impacts of climate variability and land management. Due to differing model objectives and differences in monitored data, there are currently no universally accepted procedures for model calibration and validation in the literature. In an effort to develop accepted model calibration...

  8. Calibrating the radiation detector of the ventilation of a PET radiopharmaceutical production facility

    International Nuclear Information System (INIS)

    Lacerda, Marco Aurelio de Sousa; Tavares, Jose Carlos Freitas; Silva, Juliana Batista da

    2011-01-01

    The aim of this work is to demonstrate a new methodology of estimating the calibration factor of the ventilation duct of a PET radiopharmaceutical facility. The proposed methodology was studied to minimize contamination risks for the workers, as well as the uncertainties attributed to the gas sampling. The studied facility was the Development Centre of Nuclear Technology (CDTN/CNEN) in Belo Horizonte, Brazil. It was performed 3 consecutive irradiations with normal water (H 2 16 O) for production of nitrogen-13 to estimate the calibration factor of the detector located in the chimney of the facility. The readings of the detector were registered by the online radiation monitoring system (MEDISMARTS) during the transfer of the irradiated liquid until the count rate decreased for the background (BG) levels. The remaining activity of the water from the vial was measured and the decay corrected to the beginning of the transfer of the activity. The mean calibration factor estimated was (3.6 +- 0.5) kBq . m -3 . cps -1 . The maximum activities registered in the three irradiations were, respectively, 278 s, 370 s and 366 s after transferring of the activity to the hot cell. The conservative assumptions adopted and the values found for the calibration factor, which were close to the manufacturer published data, permit to estimate, safely, the discharges of radioactive gases in the installation. (author)

  9. Calibrating the CsI(Tl) detectors of the GARFIELD apparatus

    CERN Document Server

    Abbondanno, U; Casini, G; Cavaletti, R; Cavallaro, S; Chiari, M; D'Agostino, M; Gramegna, F; Lanchais, A; Margagliotti, G V; Mastinu, P F; Milazzo, P M; Moroni, A; Nannini, A; Ordine, A; Vannini, G; Vannucci, L

    2002-01-01

    The energy and charge dependence of the light output of the CsI(Tl) detectors of the GARFIELD apparatus has been investigated for heavy ions with 5<=Z<=16 in the energy range from 2.2 to 8.3 A MeV. The results have been compared to an analytical expression successfully used in previous calibration procedures at higher energies, and a rather good agreement was obtained between measured and calculated quantities. The resulting parameter set was successfully applied to another set of experimental data. The overall result demonstrates the validity of the above mentioned calibration procedure in a wide range of incident ion energies and masses.

  10. Calibration and verification of thermographic cameras for geometric measurements

    Science.gov (United States)

    Lagüela, S.; González-Jorge, H.; Armesto, J.; Arias, P.

    2011-03-01

    Infrared thermography is a technique with an increasing degree of development and applications. Quality assessment in the measurements performed with the thermal cameras should be achieved through metrology calibration and verification. Infrared cameras acquire temperature and geometric information, although calibration and verification procedures are only usual for thermal data. Black bodies are used for these purposes. Moreover, the geometric information is important for many fields as architecture, civil engineering and industry. This work presents a calibration procedure that allows the photogrammetric restitution and a portable artefact to verify the geometric accuracy, repeatability and drift of thermographic cameras. These results allow the incorporation of this information into the quality control processes of the companies. A grid based on burning lamps is used for the geometric calibration of thermographic cameras. The artefact designed for the geometric verification consists of five delrin spheres and seven cubes of different sizes. Metrology traceability for the artefact is obtained from a coordinate measuring machine. Two sets of targets with different reflectivity are fixed to the spheres and cubes to make data processing and photogrammetric restitution possible. Reflectivity was the chosen material propriety due to the thermographic and visual cameras ability to detect it. Two thermographic cameras from Flir and Nec manufacturers, and one visible camera from Jai are calibrated, verified and compared using calibration grids and the standard artefact. The calibration system based on burning lamps shows its capability to perform the internal orientation of the thermal cameras. Verification results show repeatability better than 1 mm for all cases, being better than 0.5 mm for the visible one. As it must be expected, also accuracy appears higher in the visible camera, and the geometric comparison between thermographic cameras shows slightly better

  11. Hands-on 2.0: improving transfer of training via the Society of American Gastrointestinal and Endoscopic Surgeons (SAGES) Acquisition of Data for Outcomes and Procedure Transfer (ADOPT) program.

    Science.gov (United States)

    Dort, Jonathan; Trickey, Amber; Paige, John; Schwarz, Erin; Dunkin, Brian

    2017-08-01

    Practicing surgeons commonly learn new procedures and techniques by attending a "hands-on" course, though trainings are often ineffective at promoting subsequent procedure adoption in practice. We describe implementation of a new program with the SAGES All Things Hernia Hands-On Course, Acquisition of Data for Outcomes and Procedure Transfer (ADOPT), which employs standardized, proven teaching techniques, and 1-year mentorship. Attendee confidence and procedure adoption are compared between standard and ADOPT programs. For the pilot ADOPT course implementation, a hands-on course focusing on abdominal wall hernia repair was chosen. ADOPT participants were recruited among enrollees for the standard Hands-On Hernia Course. Enrollment in ADOPT was capped at 10 participants and limited to a 2:1 student-to-faculty ratio, compared to the standard course 22 participants with a 4:1 student-to-faculty ratio. ADOPT mentors interacted with participants through webinars, phone conferences, and continuous email availability throughout the year. All participants were asked to provide pre- and post-course surveys inquiring about the number of targeted hernia procedures performed and related confidence level. Four of 10 ADOPT participants (40%) and six of 22 standard training participants (27%) returned questionnaires. Over the 3 months following the course, ADOPT participants performed more ventral hernia mesh insertion procedures than standard training participants (median 13 vs. 0.5, p = 0.010) and considerably more total combined procedures (median 26 vs. 7, p = 0.054). Compared to standard training, learners who participated in ADOPT reported greater confidence improvements in employing a components separation via an open approach (p = 0.051), and performing an open transversus abdominis release, though the difference did not achieve statistical significance (p = 0.14). These results suggest that the ADOPT program, with standardized and structured teaching

  12. Mathematical properties of numerical inversion for jet calibrations

    Energy Technology Data Exchange (ETDEWEB)

    Cukierman, Aviv [Physics Department, Stanford University, Stanford, CA 94305 (United States); SLAC National Accelerator Laboratory, Stanford University, Menlo Park, CA 94025 (United States); Nachman, Benjamin, E-mail: bnachman@cern.ch [Physics Department, Stanford University, Stanford, CA 94305 (United States); SLAC National Accelerator Laboratory, Stanford University, Menlo Park, CA 94025 (United States); Physics Division, Lawrence Berkeley National Laboratory, Berkeley, CA 94704 (United States)

    2017-06-21

    Numerical inversion is a general detector calibration technique that is independent of the underlying spectrum. This procedure is formalized and important statistical properties are presented, using high energy jets at the Large Hadron Collider as an example setting. In particular, numerical inversion is inherently biased and common approximations to the calibrated jet energy tend to over-estimate the resolution. Analytic approximations to the closure and calibrated resolutions are demonstrated to effectively predict the full forms under realistic conditions. Finally, extensions of numerical inversion are presented which can reduce the inherent biases. These methods will be increasingly important to consider with degraded resolution at low jet energies due to a much higher instantaneous luminosity in the near future.

  13. Calibration of groundwater vulnerability mapping using the generalized reduced gradient method.

    Science.gov (United States)

    Elçi, Alper

    2017-12-01

    Groundwater vulnerability assessment studies are essential in water resources management. Overlay-and-index methods such as DRASTIC are widely used for mapping of groundwater vulnerability, however, these methods mainly suffer from a subjective selection of model parameters. The objective of this study is to introduce a calibration procedure that results in a more accurate assessment of groundwater vulnerability. The improvement of the assessment is formulated as a parameter optimization problem using an objective function that is based on the correlation between actual groundwater contamination and vulnerability index values. The non-linear optimization problem is solved with the generalized-reduced-gradient (GRG) method, which is numerical algorithm based optimization method. To demonstrate the applicability of the procedure, a vulnerability map for the Tahtali stream basin is calibrated using nitrate concentration data. The calibration procedure is easy to implement and aims the maximization of correlation between observed pollutant concentrations and groundwater vulnerability index values. The influence of each vulnerability parameter in the calculation of the vulnerability index is assessed by performing a single-parameter sensitivity analysis. Results of the sensitivity analysis show that all factors are effective on the final vulnerability index. Calibration of the vulnerability map improves the correlation between index values and measured nitrate concentrations by 19%. The regression coefficient increases from 0.280 to 0.485. It is evident that the spatial distribution and the proportions of vulnerability class areas are significantly altered with the calibration process. Although the applicability of the calibration method is demonstrated on the DRASTIC model, the applicability of the approach is not specific to a certain model and can also be easily applied to other overlay-and-index methods. Copyright © 2017 Elsevier B.V. All rights reserved.

  14. CALIPSO lidar calibration at 532 nm: version 4 nighttime algorithm

    Science.gov (United States)

    Kar, Jayanta; Vaughan, Mark A.; Lee, Kam-Pui; Tackett, Jason L.; Avery, Melody A.; Garnier, Anne; Getzewich, Brian J.; Hunt, William H.; Josset, Damien; Liu, Zhaoyan; Lucker, Patricia L.; Magill, Brian; Omar, Ali H.; Pelon, Jacques; Rogers, Raymond R.; Toth, Travis D.; Trepte, Charles R.; Vernier, Jean-Paul; Winker, David M.; Young, Stuart A.

    2018-03-01

    Data products from the Cloud-Aerosol Lidar with Orthogonal Polarization (CALIOP) on board Cloud-Aerosol Lidar and Infrared Pathfinder Satellite Observations (CALIPSO) were recently updated following the implementation of new (version 4) calibration algorithms for all of the Level 1 attenuated backscatter measurements. In this work we present the motivation for and the implementation of the version 4 nighttime 532 nm parallel channel calibration. The nighttime 532 nm calibration is the most fundamental calibration of CALIOP data, since all of CALIOP's other radiometric calibration procedures - i.e., the 532 nm daytime calibration and the 1064 nm calibrations during both nighttime and daytime - depend either directly or indirectly on the 532 nm nighttime calibration. The accuracy of the 532 nm nighttime calibration has been significantly improved by raising the molecular normalization altitude from 30-34 km to the upper possible signal acquisition range of 36-39 km to substantially reduce stratospheric aerosol contamination. Due to the greatly reduced molecular number density and consequently reduced signal-to-noise ratio (SNR) at these higher altitudes, the signal is now averaged over a larger number of samples using data from multiple adjacent granules. Additionally, an enhanced strategy for filtering the radiation-induced noise from high-energy particles was adopted. Further, the meteorological model used in the earlier versions has been replaced by the improved Modern-Era Retrospective analysis for Research and Applications, Version 2 (MERRA-2), model. An aerosol scattering ratio of 1.01 ± 0.01 is now explicitly used for the calibration altitude. These modifications lead to globally revised calibration coefficients which are, on average, 2-3 % lower than in previous data releases. Further, the new calibration procedure is shown to eliminate biases at high altitudes that were present in earlier versions and consequently leads to an improved representation of

  15. Employing an Incentive Spirometer to Calibrate Tidal Volumes Estimated from a Smartphone Camera

    Directory of Open Access Journals (Sweden)

    Bersain A. Reyes

    2016-03-01

    Full Text Available A smartphone-based tidal volume (VT estimator was recently introduced by our research group, where an Android application provides a chest movement signal whose peak-to-peak amplitude is highly correlated with reference VT measured by a spirometer. We found a Normalized Root Mean Squared Error (NRMSE of 14.998% ± 5.171% (mean ± SD when the smartphone measures were calibrated using spirometer data. However, the availability of a spirometer device for calibration is not realistic outside clinical or research environments. In order to be used by the general population on a daily basis, a simple calibration procedure not relying on specialized devices is required. In this study, we propose taking advantage of the linear correlation between smartphone measurements and VT to obtain a calibration model using information computed while the subject breathes through a commercially-available incentive spirometer (IS. Experiments were performed on twelve (N = 12 healthy subjects. In addition to corroborating findings from our previous study using a spirometer for calibration, we found that the calibration procedure using an IS resulted in a fixed bias of −0.051 L and a RMSE of 0.189 ± 0.074 L corresponding to 18.559% ± 6.579% when normalized. Although it has a small underestimation and slightly increased error, the proposed calibration procedure using an IS has the advantages of being simple, fast, and affordable. This study supports the feasibility of developing a portable smartphone-based breathing status monitor that provides information about breathing depth, in addition to the more commonly estimated respiratory rate, on a daily basis.

  16. Option price calibration from Renyi entropy

    International Nuclear Information System (INIS)

    Brody, Dorje C.; Buckley, Ian R.C.; Constantinou, Irene C.

    2007-01-01

    The calibration of the risk-neutral density function for the future asset price, based on the maximisation of the entropy measure of Renyi, is proposed. Whilst the conventional approach based on the use of logarithmic entropy measure fails to produce the observed power-law distribution when calibrated against option prices, the approach outlined here is shown to produce the desired form of the distribution. Procedures for the maximisation of the Renyi entropy under constraints are outlined in detail, and a number of interesting properties of the resulting power-law distributions are also derived. The result is applied to efficiently evaluate prices of path-independent derivatives

  17. The Chandra Source Catalog 2.0: Calibrations

    Science.gov (United States)

    Graessle, Dale E.; Evans, Ian N.; Rots, Arnold H.; Allen, Christopher E.; Anderson, Craig S.; Budynkiewicz, Jamie A.; Burke, Douglas; Chen, Judy C.; Civano, Francesca Maria; D'Abrusco, Raffaele; Doe, Stephen M.; Evans, Janet D.; Fabbiano, Giuseppina; Gibbs, Danny G., II; Glotfelty, Kenny J.; Grier, John D.; Hain, Roger; Hall, Diane M.; Harbo, Peter N.; Houck, John C.; Lauer, Jennifer L.; Laurino, Omar; Lee, Nicholas P.; Martínez-Galarza, Juan Rafael; McCollough, Michael L.; McDowell, Jonathan C.; Miller, Joseph; McLaughlin, Warren; Morgan, Douglas L.; Mossman, Amy E.; Nguyen, Dan T.; Nichols, Joy S.; Nowak, Michael A.; Paxson, Charles; Plummer, David A.; Primini, Francis Anthony; Siemiginowska, Aneta; Sundheim, Beth A.; Tibbetts, Michael; Van Stone, David W.; Zografou, Panagoula

    2018-01-01

    Among the many enhancements implemented for the release of Chandra Source Catalog (CSC) 2.0 are improvements in the processing calibration database (CalDB). We have included a thorough overhaul of the CalDB software used in the processing. The software system upgrade, called "CalDB version 4," allows for a more rational and consistent specification of flight configurations and calibration boundary conditions. Numerous improvements in the specific calibrations applied have also been added. Chandra's radiometric and detector response calibrations vary considerably with time, detector operating temperature, and position on the detector. The CalDB has been enhanced to provide the best calibrations possible to each observation over the fifteen-year period included in CSC 2.0. Calibration updates include an improved ACIS contamination model, as well as updated time-varying gain (i.e., photon energy) and quantum efficiency maps for ACIS and HRC-I. Additionally, improved corrections for the ACIS quantum efficiency losses due to CCD charge transfer inefficiency (CTI) have been added for each of the ten ACIS detectors. These CTI corrections are now time and temperature-dependent, allowing ACIS to maintain a 0.3% energy calibration accuracy over the 0.5-7.0 keV range for any ACIS source in the catalog. Radiometric calibration (effective area) accuracy is estimated at ~4% over that range. We include a few examples where improvements in the Chandra CalDB allow for improved data reduction and modeling for the new CSC.This work has been supported by NASA under contract NAS 8-03060 to the Smithsonian Astrophysical Observatory for operation of the Chandra X-ray Center.

  18. Photometric Calibration of the SPRED at the FTU Tokamak

    International Nuclear Information System (INIS)

    May, M J

    1999-01-01

    The SPRED spectrometer was photometrically calibrated by using the FTU tokamak plasma and the Grazing Incidence Time Resolving Spectrometer (GRITS) from the Johns Hopkins University [Stratton, Nucl. Fusion, Vol. 24, No. 6, pp. 767-777, 1984]. The photometric calibration of the GRITS spectrometer was transferred to the SPRED [Fonck, R.J., Applied Optics, Vol. 21, No. 12, p. 2115 (1982)] by directly comparing the intensity of bright lines emitted from the FTU tokamak plasma that were simultaneously measured by both spectrometers. The GRITS spectrometer (λ = 10 - 360 (angstrom); Δλ ∼ 0.7 (angstrom)) was photometrically calibrated in the 50 - 360 (angstrom) spectral range at the SURF II synchrotron light source at NIST in Gaithersburg MD in August 1997. The calibration of each SPRED grating was performed separately. These gratings covered the short wavelengths: 100 - 300 (angstrom)(Δλ - 1.4 (angstrom)) and the long wavelengths: 200 - 1800 (angstrom) (Δλ ∼ 7 (angstrom)). This calibration should be accurate until the microchannel plate of the SPRED is exposed to atmospheric pressure. This calibration is similar to the one obtained by Stratton [Stratton, Rev. Sci. Instrum. 57 (8), pp. 204,3 August 1986

  19. In situ neutron moisture meter calibration in lateritic soils

    International Nuclear Information System (INIS)

    Ruprecht, J.K.; Schofield, N.J.

    1990-01-01

    An in situ calibration procedure for complex lateritic soils of the jarrah forest of Western Australia is described. The calibration is based on non-destructive sampling of each access tube and on a regression of change in water content on change in neutron count ratio at 'wet' and 'dry' times of the year. Calibration equations with adequate precision were produced. However, there were high residual errors in the calibration equations which were due to a number of factors including soil water variability, the presence of a duricrust layer, soil sampling of gravelly soils and the variability of the cement slurry annulus surrounding each access tube. The calibration equations derived did not compare well with those from other studies in south-west Western Australia, but there was reasonable agreement with the general equations obtained by the Institute of Hydrology, U.K. 15 refs., 6 figs., 2 tabs

  20. Studies on dosimetric tests applying source irradiation force of Cs-137 for using in chambers for calibration and TLD type dosimeters

    International Nuclear Information System (INIS)

    Ribeiro, Laila Lorena X.; Barbosa, Rugles Cesar; Correa, Rosangela S.

    2011-01-01

    The West Central region of Brazil does not have a basic infrastructure for research, development, training programs, and personnel dosimetry education. All of them applied to environmental, industrial and medical uses. Service deployment for irradiance of TLD, via 137 Cs irradiator J. L. SHEPHERD model 28-8A (444 activity GBq) in CRCN-CO, it is necessary to introduce procedures for calibration of the radiator and other procedures related to dosimetry and calibration. Such procedures should be repeated periodically, as necessary to introduce techniques that make the service of the CRCN-CO a template, and that meet all standards requirements for radioprotection and operation of dosimetry and calibration. The objective of this work was to evaluate the radiation field of Cs-137, and the automatic system which systematizes the calibration procedures attached to a system control target for the radiator/calibration of monitors, and portable dosimeters. (author)

  1. Compensation for the Effects of Ambient Conditions on the Calibration of Multi-Capillary Pressure Drop Standards

    Directory of Open Access Journals (Sweden)

    Colard S

    2014-12-01

    Full Text Available Cigarette draw resistance and filter pressure drop (PD are both major physical parameters for the tobacco industry. Therefore these parameters must be measured reliably. For these measurements, specific equipment calibrated with PD transfer standards is used. Each transfer standard must have a known and stable PD value, such standards usually being composed of several capillary tubes associated in parallel. However, PD values are modified by ambient conditions during calibration of such standards, i.e. by temperature and relative humidity (RH of air, and atmospheric pressure. In order to reduce the influence of these ambient factors, a simplified model was developed for compensating the effects of ambient conditions on the calibration of multi-capillary PD standards.

  2. Quality control for dose calibrators

    International Nuclear Information System (INIS)

    Mendes, L.C.G.

    1984-01-01

    Nuclear medicine laboratories are required to assay samples of radioactivity to be administered to patients. Almost universally, these assays are accomplished by use of a well ionization chamber isotope calibrator. The Instituto de Radioprotecao e Dosimetria (Institute for Radiological Protection and Dosimetry) of the Comissao Nacional de Energia Nuclear (National Commission for Nuclear Energy) is carrying out a National Quality Control Programme in Nuclear Medicine, supported by the International Atomic Energy Agency. The assessment of the current needs and practices of quality control in the entire country of Brazil includes Dose Calibrators and Scintillation Cameras, but this manual is restricted to the former. Quality Control Procedures for these Instruments are described in this document together with specific recommendations and assessment of its accuracy. (author)

  3. Calibration process of highly parameterized semi-distributed hydrological model

    Science.gov (United States)

    Vidmar, Andrej; Brilly, Mitja

    2017-04-01

    Hydrological phenomena take place in the hydrological system, which is governed by nature, and are essentially stochastic. These phenomena are unique, non-recurring, and changeable across space and time. Since any river basin with its own natural characteristics and any hydrological event therein, are unique, this is a complex process that is not researched enough. Calibration is a procedure of determining the parameters of a model that are not known well enough. Input and output variables and mathematical model expressions are known, while only some parameters are unknown, which are determined by calibrating the model. The software used for hydrological modelling nowadays is equipped with sophisticated algorithms for calibration purposes without possibility to manage process by modeler. The results are not the best. We develop procedure for expert driven process of calibration. We use HBV-light-CLI hydrological model which has command line interface and coupling it with PEST. PEST is parameter estimation tool which is used widely in ground water modeling and can be used also on surface waters. Process of calibration managed by expert directly, and proportionally to the expert knowledge, affects the outcome of the inversion procedure and achieves better results than if the procedure had been left to the selected optimization algorithm. First step is to properly define spatial characteristic and structural design of semi-distributed model including all morphological and hydrological phenomena, like karstic area, alluvial area and forest area. This step includes and requires geological, meteorological, hydraulic and hydrological knowledge of modeler. Second step is to set initial parameter values at their preferred values based on expert knowledge. In this step we also define all parameter and observation groups. Peak data are essential in process of calibration if we are mainly interested in flood events. Each Sub Catchment in the model has own observations group

  4. Calibration and consistency of results of an ionization-chamber secondary standard measuring system for activity

    International Nuclear Information System (INIS)

    Schrader, Heinrich

    2000-01-01

    Calibration in terms of activity of the ionization-chamber secondary standard measuring systems at the PTB is described. The measurement results of a Centronic IG12/A20, a Vinten ISOCAL IV and a radionuclide calibrator chamber for nuclear medicine applications are discussed, their energy-dependent efficiency curves established and the consistency checked using recently evaluated radionuclide decay data. Criteria for evaluating and transferring calibration factors (or efficiencies) are given

  5. Emissivity-corrected power loss calibration for lock-in thermography measurements on silicon solar cells

    International Nuclear Information System (INIS)

    Kasemann, Martin; Walter, Benjamin; Meinhardt, Christoph; Ebser, Jan; Kwapil, Wolfram; Warta, Wilhelm

    2008-01-01

    This paper describes power loss calibration procedures with implemented emissivity correction. The determination of our emissivity correction matrix does neither rely on blackbody reference measurements nor on the knowledge of any sample temperatures. To describe the emissivity-corrected power calibration procedures in detail, we review the theory behind lock-in thermography and show experimentally that the lock-in signal is proportional to the power dissipation in the solar cell. Experiments show the successful application of our emissivity correction procedure, which significantly improves the informative value of lock-in thermography images and the reliability of the conclusions drawn from these images

  6. Calibration and monitoring of a scintillator HCAL with SiPMs CALICE scintillator HCAL

    International Nuclear Information System (INIS)

    Lucaci-Timoce, Angela

    2009-01-01

    The operational experience with a highly-granular analogue hadronic calorimeter (AHCAL) consisting of 7608 individual scintillator tiles readout via Silicon-Photo-multipliers (SiPM) is presented. The calibration of each cell is based on minimum ionizing particle signals for which in general a muon beam is used. In addition, a correction for the non-linearity introduced by the finite number of pixels (1156) in the SiPM is applied. The aspects of temperature and voltage dependence of SiPM are addressed, and monitoring and calibration procedures are discussed. Such procedures are essential for the extrapolation of calibration factors over several days of data taking with the calorimeter. For this purpose a versatile UV-LED light distribution system was developed, capable of delivering light to all tiles with intensity from a few photo-electrons to the saturation of the SiPM. The procedures are tested using data collected with the AHCAL at the CERN SPS test beam.

  7. Electronographic calibration of UK 1.2-m Schmidt plates

    International Nuclear Information System (INIS)

    Hawkins, M.R.S.

    1979-01-01

    Two electronographic sequences are given in the South Galactic Pole region down to msub(B) = approximately 23 +- 0.3 mag. These sequences are used to obtain a calibration for COSMOS measures of UK 1.2-m Schmidt plates and evaluate their photometric transfer properties. (author)

  8. D0 Silicon Upgrade: End Calorimeter Transfer Bridge Design and Installation Procedures

    International Nuclear Information System (INIS)

    Stredde, H.J.

    1991-01-01

    elevated to the level of the transfer bridge. The elevation process is the same as that used for the CC installation. When the correct elevation is attained the transporter extensions are positioned and fastened to the transporter on the south side. Static column supports will be added underneath the transporter to minimize deflections as the NEC is rolled off. The carriage extensions are now put in place and the Hilman rollers transferred into position under the extensions. The NEC is now ready to be pulled off of the transporter and onto the transfer bridge. Again using the Cyl. and chain mechanism, with the anchor point attached to the transporter extension on the north sidewalk, pull the NEC completely off of the transporter and onto the transfer bridge. Once the NEC is on the transfer bridge a rearrangement of equipment must be done. The transporter extension on the north sidewalk is decoupled from the transfer bridge and moved against the wall of the fixed counting area. The transporter extension at the south end of the transporter is removed, the transporter is decoupled from and moved to the north side of the transfer bridge. The transporter is supported from the floor at an elevation equal to the transfer bridge and coupled to it. The transporter extension is fastened to the north end of the transporter. The movement of the NEC now continues until it is back on the transporter. At this time the transfer bridge, concrete shield blocks and steel plates in the pit are removed. The detector is moved back to its normal parked position(cable bridge up). The NEC is lowered to the sidewalk (reverse of elevation procedure). Welding of the heads is then finished prior to final installation on the center beam. The final installation is identical to the CC installation, using all the same equipment. The finite element analysis of the transfer bridge design was done by Kay Weber of the Accelerator Mechanical Support Group.

  9. ATLAS Tile calorimeter calibration and monitoring systems

    CERN Document Server

    Boumediene, Djamel Eddine; The ATLAS collaboration

    2017-01-01

    The ATLAS Tile Calorimeter (TileCal) is the central section of the hadronic calorimeter of the ATLAS experiment and provides important information for reconstruction of hadrons, jets, hadronic decays of tau leptons and missing transverse energy. This sampling calorimeter uses steel plates as absorber and scintillating tiles as active medium. The light produced by the passage of charged particles is transmitted by wavelength shifting fibres to photomultiplier tubes (PMTs). PMT signals are then digitized at 40 MHz and stored on detector and are only transferred off detector once the first level trigger acceptance has been confirmed. The readout is segmented into about 5000 cells (longitudinally and transversally), each of them being read out by two PMTs in parallel. To calibrate and monitor the stability and performance of each part of the readout chain, a set of calibration systems is used. The TileCal calibration system comprises Cesium radioactive sources, laser, charge injection elements and an integrator b...

  10. Calibration of dosemeters and survey instruments for photon

    International Nuclear Information System (INIS)

    Alamares, A.L.; Caseria, E.S.

    1995-01-01

    The Philippine Nuclear Research Institute (PNRI), the country's regulatory agency, through its Radiation Protection Unit of the Nuclear Regulations Licensing and Safeguard Division provides the calibration services of nuclear instruments used by various institutions. The periodic calibration and standardization of radiation and protection survey instruments are done to insure correct and valid radiation readings. It is also a regulatory requirement for radiation workers to use only operable and precalibrated survey instruments for their work with radioactive materials. The PNRI maintains and operates the Secondary Standard Dosimetry Laboratory (SSDL). It provides the periodic calibration and standardization of radiation monitoring and protection survey instruments, and also the radiological hazard and performance evaluation surveys of radiotherapy facilities and others. For the calibration of survey instruments, the standard sources of Cs-137 and Co-60 are used. The calibration procedures for survey meters and pocket dosimeters are explained. The clients served are shown. The personnel dosimetry using film badges, the application of ICRU operational quantities, and the problems encountered and recommendations are reported. (K.I.)

  11. Onboard calibration and monitoring for the SWIFT instrument

    International Nuclear Information System (INIS)

    Rahnama, P; McDade, I; Shepherd, G; Gault, W

    2012-01-01

    The SWIFT (Stratospheric Wind Interferometer for Transport studies) instrument is a proposed space-based field-widened Doppler Michelson interferometer designed to measure stratospheric winds and ozone densities using a passive optical technique called Doppler Michelson imaging interferometry. The onboard calibration and monitoring procedures for the SWIFT instrument are described in this paper. Sample results of the simulations of onboard calibration measurements are presented and discussed. This paper also discusses the results of the derivation of the calibrations and monitoring requirements for the SWIFT instrument. SWIFT's measurement technique and viewing geometry are briefly described. The reference phase calibration and filter monitoring for the SWIFT instrument are two of the main critical design issues. In this paper it is shown that in order to meet SWIFT's science requirements, Michelson interferometer optical path difference monitoring corresponding to a phase calibration accuracy of ∼10 −3 radians, filter passband monitoring corresponding to phase accuracy of ∼5 × 10 −3 radians and a thermal stability of 10 −3 K s −1 are required. (paper)

  12. Laser calibration of the ATLAS Tile Calorimeter

    CERN Document Server

    Di Gregorio, Giulia; The ATLAS collaboration

    2017-01-01

    High performance stability of the ATLAS Tile calorimeter is achieved with a set of calibration procedures. One step of the calibrtion procedure is based on measurements of the response stability to laser excitation of the photomultipliers (PMTs) that are used to readout the calorimeter cells. A facility to study in lab the PMT stability response is operating in the PISA-INFN laboratories since 2015. Goals of the test in lab are to study the time evolution of the PMT response to reproduce and to understand the origin of the resonse drifts seen with the PMT mounted on the Tile calorimeter in its normal operation during LHC run I and run II. A new statistical approach was developed to measure the drift of the absolute gain. This approach was applied to both the ATLAS laser calibration data and to the data collected in the Pisa local laboratory. The preliminary results from these two studies are shown.

  13. Towards a global network of gamma-ray detector calibration facilities

    Science.gov (United States)

    Tijs, Marco; Koomans, Ronald; Limburg, Han

    2016-09-01

    Gamma-ray logging tools are applied worldwide. At various locations, calibration facilities are used to calibrate these gamma-ray logging systems. Several attempts have been made to cross-correlate well known calibration pits, but this cross-correlation does not include calibration facilities in Europe or private company calibration facilities. Our aim is to set-up a framework that gives the possibility to interlink all calibration facilities worldwide by using `tools of opportunity' - tools that have been calibrated in different calibration facilities, whether this usage was on a coordinated basis or by coincidence. To compare the measurement of different tools, it is important to understand the behaviour of the tools in the different calibration pits. Borehole properties, such as diameter, fluid, casing and probe diameter strongly influence the outcome of gamma-ray borehole logging. Logs need to be properly calibrated and compensated for these borehole properties in order to obtain in-situ grades or to do cross-hole correlation. Some tool providers provide tool-specific correction curves for this purpose. Others rely on reference measurements against sources of known radionuclide concentration and geometry. In this article, we present an attempt to set-up a framework for transferring `local' calibrations to be applied `globally'. This framework includes corrections for any geometry and detector size to give absolute concentrations of radionuclides from borehole measurements. This model is used to compare measurements in the calibration pits of Grand Junction, located in the USA; Adelaide (previously known as AMDEL), located in Adelaide Australia; and Stonehenge, located at Medusa Explorations BV in the Netherlands.

  14. Calibration of neutrons monitors with moderators and application in the calibration factors of albedo dosemeters

    International Nuclear Information System (INIS)

    Schuch, L.A.

    1978-11-01

    The calibration factors and the reproducibility of an Albedo Dosimeter designed for personal neutron monitoring were determined. These factor were obtained simulating the dosimeter reading and the equivalent dose in the locality by a convenient combination of responses of the Bonner Sphere Spectrometer. The results obtained in the simulation were verified experimentally for different spectra employing the Am-Be, bare 252 Cf source and 253 Cf source with graphite sields of varying thickness. Different standards were used in the procedures necessary for the determination of the calibration factors. An Am-Be neutron source, standardized by the activation of a manganese sulphate bath was used as a primary standard. As a secondary standard, for the measurement of the neutron fluence, a De Pangher Long Counter was used and the scattering effects were determined using the shadow cone method. The other monitors such as the Rem-Counter and the Bonner Sphere Spectrometer were also calibrated with reference to the secondary standard with a view to comparing the results obtained with those furnished by the Albedo Dosimeter. (Author) [pt

  15. Binary pseudo-random patterned structures for modulation transfer function calibration and resolution characterization of a full-field transmission soft x-ray microscope

    Energy Technology Data Exchange (ETDEWEB)

    Yashchuk, V. V., E-mail: VVYashchuk@lbl.gov; Chan, E. R.; Lacey, I. [Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, California 94720 (United States); Fischer, P. J. [Center for X-Ray Optics, Lawrence Berkeley National Laboratory, Berkeley, California 94720 (United States); Physics Department, University of California Santa Cruz, Santa Cruz, California 94056 (United States); Conley, R. [Advance Photon Source, Argonne National Laboratory, Argonne, Illinois 60439 (United States); National Synchrotron Light Source II, Brookhaven National Laboratory, Upton, New York 11973 (United States); McKinney, W. R. [Diablo Valley College, 321 Golf Club Road, Pleasant Hill, California 94523 (United States); Artemiev, N. A. [KLA-Tencor Corp., 1 Technology Drive, Milpitas, California 95035 (United States); Bouet, N. [National Synchrotron Light Source II, Brookhaven National Laboratory, Upton, New York 11973 (United States); Cabrini, S. [Molecular Foundry, Lawrence Berkeley National Laboratory, Berkeley, California 94720 (United States); Calafiore, G.; Peroz, C.; Babin, S. [aBeam Technologies, Inc., Hayward, California 94541 (United States)

    2015-12-15

    We present a modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) one-dimensional sequences and two-dimensional arrays as an effective method for spectral characterization in the spatial frequency domain of a broad variety of metrology instrumentation, including interferometric microscopes, scatterometers, phase shifting Fizeau interferometers, scanning and transmission electron microscopes, and at this time, x-ray microscopes. The inherent power spectral density of BPR gratings and arrays, which has a deterministic white-noise-like character, allows a direct determination of the MTF with a uniform sensitivity over the entire spatial frequency range and field of view of an instrument. We demonstrate the MTF calibration and resolution characterization over the full field of a transmission soft x-ray microscope using a BPR multilayer (ML) test sample with 2.8 nm fundamental layer thickness. We show that beyond providing a direct measurement of the microscope’s MTF, tests with the BPRML sample can be used to fine tune the instrument’s focal distance. Our results confirm the universality of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters.

  16. Binary pseudo-random patterned structures for modulation transfer function calibration and resolution characterization of a full-field transmission soft x-ray microscope

    International Nuclear Information System (INIS)

    Yashchuk, V. V.; Chan, E. R.; Lacey, I.; Fischer, P. J.; Conley, R.; McKinney, W. R.; Artemiev, N. A.; Bouet, N.; Cabrini, S.; Calafiore, G.; Peroz, C.; Babin, S.

    2015-01-01

    We present a modulation transfer function (MTF) calibration method based on binary pseudo-random (BPR) one-dimensional sequences and two-dimensional arrays as an effective method for spectral characterization in the spatial frequency domain of a broad variety of metrology instrumentation, including interferometric microscopes, scatterometers, phase shifting Fizeau interferometers, scanning and transmission electron microscopes, and at this time, x-ray microscopes. The inherent power spectral density of BPR gratings and arrays, which has a deterministic white-noise-like character, allows a direct determination of the MTF with a uniform sensitivity over the entire spatial frequency range and field of view of an instrument. We demonstrate the MTF calibration and resolution characterization over the full field of a transmission soft x-ray microscope using a BPR multilayer (ML) test sample with 2.8 nm fundamental layer thickness. We show that beyond providing a direct measurement of the microscope’s MTF, tests with the BPRML sample can be used to fine tune the instrument’s focal distance. Our results confirm the universality of the method that makes it applicable to a large variety of metrology instrumentation with spatial wavelength bandwidths from a few nanometers to hundreds of millimeters

  17. Intercomparison of calibration procedures of high dose rate {sup 192} Ir sources in Brazil and a proposal of a new methodology; Intercomparacao de procedimientos de calibracao de fontes de {sup 192} Ir de alta taxa de dose no Brasil e proposta de uma nova metodologia

    Energy Technology Data Exchange (ETDEWEB)

    Marechal, M.H.; Almeida, C.E. de [Laboratorio Nacional de Metrologia das Radiacoes Ionizantes IRD/CNEN. Caixa Postal 37750 CEP 22780-160 Rio de Janeiro (Brazil)

    1998-12-31

    The objective of this paper is to report the results of an intercomparison of the calibration procedures for {sup 192} Ir sources presently in use in Brazil and to proposal a calibration procedure to derive the N{sub k} for a Farmer type ionization chamber for {sup 192} Ir energy by interpolating from a {sup 60} Co gamma-rays and 250 kV x-rays calibration factors. the intercomparison results were all within {+-} 3.0 % except one case where 4.6 % was observed and latter identified as a problem with N-k value for X-rays. The method proposed by the present work make possible the improvement of the metrological coherence among the calibration laboratories and their users once the N{sub k} values could then provided by any of the members of SSDL network. (Author)

  18. Generic System for Remote Testing and Calibration of Measuring Instruments: Security Architecture

    Science.gov (United States)

    Jurčević, M.; Hegeduš, H.; Golub, M.

    2010-01-01

    Testing and calibration of laboratory instruments and reference standards is a routine activity and is a resource and time consuming process. Since many of the modern instruments include some communication interfaces, it is possible to create a remote calibration system. This approach addresses a wide range of possible applications and permits to drive a number of different devices. On the other hand, remote calibration process involves a number of security issues due to recommendations specified in standard ISO/IEC 17025, since it is not under total control of the calibration laboratory personnel who will sign the calibration certificate. This approach implies that the traceability and integrity of the calibration process directly depends on the collected measurement data. The reliable and secure remote control and monitoring of instruments is a crucial aspect of internet-enabled calibration procedure.

  19. Quality control procedures in positron tomography

    International Nuclear Information System (INIS)

    Spinks, T.; Jones, T.; Heather, J.; Gilardi, M.

    1989-01-01

    The derivation of physiological parameters in positron tomography relies on accurate calibration of the tomograph. Normally, the calibration relates image pixel count density to the count rate from an external blood counter per unit activity concentration in each device. The quality control of the latter is simple and relies on detector stability assessed by measurement of a standard source of similar geometry to a blood sample. The quality control of the tomographic data depends on (i) detector stability, (ii) uniformity of calibration and normalisation sources and (iii) reproducibility of the attenuation correction procedure. A quality control procedure has been developed for an 8 detector ring (15 transaxial plane) tomograph in which detector response is assessed by acquiring data from retractable transmission ring sources. These are scanned daily and a print out of detector efficiencies is produced as well as changes from a given data. This provides the raw data from which decisions on recalibration or renormalisation are made. (orig.)

  20. Independent System Calibration of Sentinel-1B

    Directory of Open Access Journals (Sweden)

    Marco Schwerdt

    2017-05-01

    Full Text Available Sentinel-1B is the second of two C-Band Synthetic Aperture Radar (SAR satellites of the Sentinel-1 mission, launched in April 2016—two years after the launch of the first satellite, Sentinel-1A. In addition to the commissioning of Sentinel-1B executed by the European Space Agency (ESA, an independent system calibration was performed by the German Aerospace Center (DLR on behalf of ESA. Based on an efficient calibration strategy and the different calibration procedures already developed and applied for Sentinel-1A, extensive measurement campaigns were executed by initializing and aligning DLR’s reference targets deployed on the ground. This paper describes the different activities performed by DLR during the Sentinel-1B commissioning phase and presents the results derived from the analysis and the evaluation of a multitude of data takes and measurements.

  1. Probabilistic calibration of safety coefficients for flawed components in nuclear engineering

    International Nuclear Information System (INIS)

    Ardillon, E.; Pitner, P.; Barthelet, B.; Remond, A.

    1996-01-01

    The rules that are currently under application to verify the acceptance of flaws in nuclear components rely on deterministic criteria supposed to ensure the safe operating of plants. The interest of having a precise and reliable method to evaluate the safety margins and the integrity of components led Electricite de France to launch an approach to link directly safety coefficients with safety levels. This paper presents a probabilistic methodology to calibrate safety coefficients in relation to reliability target values. The proposed calibration procedure applies to the case of a ferritic flawed pipe using the R6 procedure for assessing the integrity of the structure. (authors). 5 refs., 5 figs

  2. Probabilistic calibration of safety coefficients for flawed components in nuclear engineering

    International Nuclear Information System (INIS)

    Ardillon, E.; Pitner, P.; Barthelet, B.; Remond, A.

    1995-01-01

    The current rules applied to verify the flaws acceptance in nuclear components rely on deterministic criteria supposed to ensure the plant safe operation. The interest in have a precise and reliable method to evaluate the safety margins and the integrity of components led Electricite de France to launch an approach to link directly safety coefficients with safety levels. This paper presents a probabilistic methodology to calibrate safety coefficients in relation do reliability target values. The proposed calibration procedure applies to the case of a ferritic flawed pipe using the R 6 procedure for assessing the structure integrity. (author). 5 refs., 5 figs., 1 tab

  3. AVIRIS Inflight Calibration Experiments, Analysis, and Results in 2000

    Science.gov (United States)

    Green, Robert O.; Pavri, Betina

    2001-01-01

    The NASA Airborne Visible/Infrared Imaging Spectrometer (AVIRIS) measures spectra from 370 to 2500 nm with nominally 10-nm sampling and resolution. The spectra are acquired as images with an 11 km width and up to 800 km length from the ER-2 platform or 2.1 km width and 160 km length from the Twin Otter platform. AVIRIS measurements are used for a range of Earth science research and application objectives. The molecular absorption and particle scattering properties expressed in the calibrated AVIRIS measurements are used. For both science research and application objectives, calibration of the AVIRIS spectra is required to produce useful results. Each year prior to the flight season AVIRIS is calibrated in the laboratory. However, the temperature, pressure, vibration, and observational geometry, as well as mechanical, electrical, and operational interfaces of the laboratory are different than the environment on board the airborne platform. To validate the calibration of AVIRIS in the flight environment, an inflight calibration experiment is orchestrated at the beginning of each flight season. In most years additional inflight calibration experiments occur towards the middle and end of the flight season as well. For an inflight calibration experiment, AVIRIS acquires airborne data over a designated calibration target. In concert with the airborne data acquisition, surface and atmospheric properties at the calibration target are measured in situ. These in situ measurements are used to constrain a radiative transfer code and predict the radiance incident at the AVIRIS instrument from the calibration target. This prediction is compared with the AVIRIS-measured radiance to validate the calibration of AVIRIS in the flight environment. Additional properties (such as the AVIRIS inflight measurement precision) are determined as well. This paper presents measurements, analyses, and results from the inflight calibration experiment held on the dry lake bed surface of Rogers Dry

  4. PLEIADES ABSOLUTE CALIBRATION : INFLIGHT CALIBRATION SITES AND METHODOLOGY

    Directory of Open Access Journals (Sweden)

    S. Lachérade

    2012-07-01

    Full Text Available In-flight calibration of space sensors once in orbit is a decisive step to be able to fulfil the mission objectives. This article presents the methods of the in-flight absolute calibration processed during the commissioning phase. Four In-flight calibration methods are used: absolute calibration, cross-calibration with reference sensors such as PARASOL or MERIS, multi-temporal monitoring and inter-bands calibration. These algorithms are based on acquisitions over natural targets such as African deserts, Antarctic sites, La Crau (Automatic calibration station and Oceans (Calibration over molecular scattering or also new extra-terrestrial sites such as the Moon and selected stars. After an overview of the instrument and a description of the calibration sites, it is pointed out how each method is able to address one or several aspects of the calibration. We focus on how these methods complete each other in their operational use, and how they help building a coherent set of information that addresses all aspects of in-orbit calibration. Finally, we present the perspectives that the high level of agility of PLEIADES offers for the improvement of its calibration and a better characterization of the calibration sites.

  5. Calibration Lessons Learned from Hyperion Experience

    Science.gov (United States)

    Casement, S.; Ho, K.; Sandor-Leahy, S.; Biggar, S.; Czapla-Myers, J.; McCorkel, J.; Thome, K.

    2009-12-01

    The use of hyperspectral imagers to provide climate-quality data sets, such as those expected from the solar reflective sensor on the Climate Absolute Radiance and Refractivity Observatory (CLARREO), requires stringent radiometric calibration requirements. These stringent requirements have been nearly met with broadband radiometers such as CERES, but high resolution spectrometers pose additional challenges. A review of the calibration processes for past space-based HSIs provide guidance on the calibration processes that will be needed for future sensors. In November 2000, the Earth Observer-1 (EO-1) platform was launched onboard a Boeing Delta II launch vehicle. The primary purpose of the EO-1 mission was to provide a technological testbed for spaceborne components. The platform has three sensors onboard, of which, the hyperspectral imager (HSI) Hyperion, is discussed here. The Hyperion sensor at the time had no comparable sensor in earth orbit, being the first grating-based, hyperspectral, civilian sensor in earth orbit. Ground and on-orbit calibration procedures including all cross-calibration activities have achieved an estimated instrument absolute radiometric error of 2.9% in the Visible channel (0.4 - 1.0 microns) and 3.4% in the shortwave infrared (SWIR, 0.9 - 2.5 microns) channel (EO-1/Hyperion Early Orbit Checkout Report Part II On-Orbit Performance Verification and Calibration). This paper describes the key components of the Hyperion calibration process that are applicable to future HSI missions. The pre-launch methods relied on then newly-developed, detector-based methods. Subsequent vicarious methods including cross-calibration with other sensors and the reflectance-based method showed significant differences from the prelaunch calibration. Such a difference demonstrated the importance of the vicarious methods as well as pointing to areas for improvement in the prelaunch methods. We also identify areas where lessons learned from Hyperion regarding

  6. Calibration of Cryogenic Thermometers for the LHC

    CERN Document Server

    Balle, Ch; Vauthier, N; Thermeau, JP

    2008-01-01

    6000 cryogenic temperature sensors of resistive type covering the range from room temperature down to 1.6 K are installed on the LHC machine. In order to meet the stringent requirements on temperature control of the superconducting magnets, each single sensor needs to be calibrated individually. In the framework of a special contribution, IPN (Institut de Physique Nucléaire) in Orsay, France built and operated a calibration facility with a throughput of 80 thermometers per week. After reception from the manufacturer, the thermometer is first assembled onto a support specific to the measurement environment, and then thermally cycled ten times and calibrated at least once from 1.6 to 300 K. The procedure for each of these interventions includes various measurements and the acquired data is recorded in an ORACLE®-database. Furthermore random calibrations on some samples are executed at CERN to crosscheck the coherence between the approximation data obtained by both IPN and CERN. In the range of 1.5 K to 30 K...

  7. New method of assigning uncertainty in volume calibration

    International Nuclear Information System (INIS)

    Lechner, J.A.; Reeve, C.P.; Spiegelman, C.H.

    1980-12-01

    This paper presents a practical statistical overview of the pressure-volume calibration curve for large nuclear materials processing tanks. It explains the appropriateness of applying splines (piecewise polynomials) to this curve, and it presents an overview of the associated statistical uncertainties. In order to implement these procedures, a practical and portable FORTRAN IV program is provided along with its users' manual. Finally, the recommended procedure is demonstrated on actual tank data collected by NBS

  8. Validation of the ATLAS hadronic calibration with the LAr End-Cap beam tests data

    International Nuclear Information System (INIS)

    Barillari, Teresa

    2009-01-01

    The high granularity of the ATLAS calorimeter and the large number of expected particles per event require a clustering algorithm that is able to suppress noise and pile-up efficiently. Therefore the cluster reconstruction is the essential first step in the hadronic calibration. The identification of electromagnetic components within a hadronic cluster using cluster shape variables is the next step in the hadronic calibration procedure. Finally the energy density of individual cells is used to assign the proper weight to correct for the invisible energy deposits of hadrons due to the non-compensating nature of the ATLAS calorimeter and to correct for energy losses in material non instrumented with read-out. The weighting scheme employs the energy density in individual cells. Therefore the validation of the Monte Carlo simulation, which is used to define the weighting parameters and energy correction algorithms, is an essential step in the hadronic calibration procedure. Pion data, obtained in a beam test corresponding to the pseudorapidity region 2.5 < |η| < 4.0 in ATLAS and in the energy range 40 GeV ≤ E ≤ 200 GeV, have been compared with Monte Carlo simulations, using the full ATLAS hadronic calibration procedure.

  9. MT3DMS: Model use, calibration, and validation

    Science.gov (United States)

    Zheng, C.; Hill, Mary C.; Cao, G.; Ma, R.

    2012-01-01

    MT3DMS is a three-dimensional multi-species solute transport model for solving advection, dispersion, and chemical reactions of contaminants in saturated groundwater flow systems. MT3DMS interfaces directly with the U.S. Geological Survey finite-difference groundwater flow model MODFLOW for the flow solution and supports the hydrologic and discretization features of MODFLOW. MT3DMS contains multiple transport solution techniques in one code, which can often be important, including in model calibration. Since its first release in 1990 as MT3D for single-species mass transport modeling, MT3DMS has been widely used in research projects and practical field applications. This article provides a brief introduction to MT3DMS and presents recommendations about calibration and validation procedures for field applications of MT3DMS. The examples presented suggest the need to consider alternative processes as models are calibrated and suggest opportunities and difficulties associated with using groundwater age in transport model calibration.

  10. Calibration of NASA Turbulent Air Motion Measurement System

    Science.gov (United States)

    Barrick, John D. W.; Ritter, John A.; Watson, Catherine E.; Wynkoop, Mark W.; Quinn, John K.; Norfolk, Daniel R.

    1996-01-01

    A turbulent air motion measurement system (TAMMS) was integrated onboard the Lockheed 188 Electra airplane (designated NASA 429) based at the Wallops Flight Facility in support of the NASA role in global tropospheric research. The system provides air motion and turbulence measurements from an airborne platform which is capable of sampling tropospheric and planetary boundary-layer conditions. TAMMS consists of a gust probe with free-rotating vanes mounted on a 3.7-m epoxy-graphite composite nose boom, a high-resolution inertial navigation system (INS), and data acquisition system. A variation of the tower flyby method augmented with radar tracking was implemented for the calibration of static pressure position error and air temperature probe. Additional flight calibration maneuvers were performed remote from the tower in homogeneous atmospheric conditions. System hardware and instrumentation are described and the calibration procedures discussed. Calibration and flight results are presented to illustrate the overall ability of the system to determine the three-component ambient wind fields during straight and level flight conditions.

  11. Novel Real-time Alignment and Calibration of the LHCb detector in Run2

    Science.gov (United States)

    Martinelli, Maurizio; LHCb Collaboration

    2017-10-01

    LHCb has introduced a novel real-time detector alignment and calibration strategy for LHC Run2. Data collected at the start of the fill are processed in a few minutes and used to update the alignment parameters, while the calibration constants are evaluated for each run. This procedure improves the quality of the online reconstruction. For example, the vertex locator is retracted and reinserted for stable beam conditions in each fill to be centred on the primary vertex position in the transverse plane. Consequently its position changes on a fill-by-fill basis. Critically, this new real-time alignment and calibration procedure allows identical constants to be used in the online and offline reconstruction, thus improving the correlation between triggered and offline-selected events. This offers the opportunity to optimise the event selection in the trigger by applying stronger constraints. The required computing time constraints are met thanks to a new dedicated framework using the multi-core farm infrastructure for the trigger. The motivation for a real-time alignment and calibration of the LHCb detector is discussed from both the operational and physics performance points of view. Specific challenges of this novel configuration are discussed, as well as the working procedures of the framework and its performance.

  12. Camera calibration method of binocular stereo vision based on OpenCV

    Science.gov (United States)

    Zhong, Wanzhen; Dong, Xiaona

    2015-10-01

    Camera calibration, an important part of the binocular stereo vision research, is the essential foundation of 3D reconstruction of the spatial object. In this paper, the camera calibration method based on OpenCV (open source computer vision library) is submitted to make the process better as a result of obtaining higher precision and efficiency. First, the camera model in OpenCV and an algorithm of camera calibration are presented, especially considering the influence of camera lens radial distortion and decentering distortion. Then, camera calibration procedure is designed to compute those parameters of camera and calculate calibration errors. High-accurate profile extraction algorithm and a checkboard with 48 corners have also been used in this part. Finally, results of calibration program are presented, demonstrating the high efficiency and accuracy of the proposed approach. The results can reach the requirement of robot binocular stereo vision.

  13. The Impact of Indoor and Outdoor Radiometer Calibration on Solar Measurements: Preprint

    Energy Technology Data Exchange (ETDEWEB)

    Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Reda, Ibrahim; Robinson, Justin

    2016-07-01

    Accurate solar radiation data sets are critical to reducing the expenses associated with mitigating performance risk for solar energy conversion systems, and they help utility planners and grid system operators understand the impacts of solar resource variability. The accuracy of solar radiation measured by radiometers depends on the instrument performance specification, installation method, calibration procedure, measurement conditions, maintenance practices, location, and environmental conditions. This study addresses the effect of calibration methodologies and the resulting calibration responsivities provided by radiometric calibration service providers such as the National Renewable Energy Laboratory (NREL) and manufacturers of radiometers. Some of these radiometers are calibrated indoors, and some are calibrated outdoors. To establish or understand the differences in calibration methodology, we processed and analyzed field-measured data from these radiometers. This study investigates calibration responsivities provided by NREL's broadband outdoor radiometer calibration (BORCAL) and a few prominent manufacturers. The reference radiometer calibrations are traceable to the World Radiometric Reference. These different methods of calibration demonstrated 1% to 2% differences in solar irradiance measurement. Analyzing these values will ultimately assist in determining the uncertainties of the radiometer data and will assist in developing consensus on a standard for calibration.

  14. 49 CFR 193.2513 - Transfer procedures.

    Science.gov (United States)

    2010-10-01

    ... ADMINISTRATION, DEPARTMENT OF TRANSPORTATION (CONTINUED) PIPELINE SAFETY LIQUEFIED NATURAL GAS FACILITIES... Practice.” (2) Before transfer, verify that each receiving container or tank vehicle does not contain any... volume of each receiving container or tank vehicle to ensure that expansion of the incoming fluid due to...

  15. CryoSat/SIRAL Cal1 Calibration Orbits

    Science.gov (United States)

    Scagliola, Michele; Fornari, Marco; Bouffard, Jerome; Parrinello, Tommaso

    2017-04-01

    The main payload of CryoSat is a Ku band pulsewidth limited radar altimeter, called SIRAL (Synthetic interferometric radar altimeter), that transmits pulses at a high pulse repetition frequency thus making the received echoes phase coherent and suitable for SAR processing. This allows to reach an along track resolution that is significantly improved with respect to traditional pulse-width limited altimeters. Due to the fact that SIRAL is a phase coherent pulse-width limited radar altimeter, a proper calibration approach has been developed. In fact, not only corrections for transfer function, gain and instrument path delay have to be computed (as in previous altimeters), but also corrections for phase (SAR/SARIn) and phase difference between the two receiving chains (SARIN only). Recalling that the CryoSat's orbit has a high inclination of 92° and it is non-sun-synchronous, the temperature of the SIRAL changes continuously along the orbit with a period of about 480 days and it is also function of the ascending/descending passes. By analysis of the CAL1 calibration corrections, it has been verified that the internal path delay and the instrument gain variation measured on the SIRAL are affected by the thermal status of the instrument and as a consequence they are expected to vary along the orbit. In order to gain knowledge on the calibration corrections (i.e. the instrument behavior) as function of latitude and temperature, it has been planned to command a few number of orbits where only CAL1 calibration acquisitions are continuously performed. The analysis of the CAL1 calibration corrections produced along the Calibration orbits can be also useful to verify whether the current calibration plan is able to provide sufficiently accurate corrections for the instrument acquisitions at any latitude. In 2016, the CryoSat/SIRAL Cal1 Calibration Orbits have been commanded two times, a first time the 20th of July 2016 and a second time the 24th of November 2016, and they

  16. Spectral Transfer Learning Using Information Geometry for a User-Independent Brain-Computer Interface.

    Science.gov (United States)

    Waytowich, Nicholas R; Lawhern, Vernon J; Bohannon, Addison W; Ball, Kenneth R; Lance, Brent J

    2016-01-01

    Recent advances in signal processing and machine learning techniques have enabled the application of Brain-Computer Interface (BCI) technologies to fields such as medicine, industry, and recreation; however, BCIs still suffer from the requirement of frequent calibration sessions due to the intra- and inter-individual variability of brain-signals, which makes calibration suppression through transfer learning an area of increasing interest for the development of practical BCI systems. In this paper, we present an unsupervised transfer method (spectral transfer using information geometry, STIG), which ranks and combines unlabeled predictions from an ensemble of information geometry classifiers built on data from individual training subjects. The STIG method is validated in both off-line and real-time feedback analysis during a rapid serial visual presentation task (RSVP). For detection of single-trial, event-related potentials (ERPs), the proposed method can significantly outperform existing calibration-free techniques as well as outperform traditional within-subject calibration techniques when limited data is available. This method demonstrates that unsupervised transfer learning for single-trial detection in ERP-based BCIs can be achieved without the requirement of costly training data, representing a step-forward in the overall goal of achieving a practical user-independent BCI system.

  17. Jurisprudential aspects regarding the action in annulment of the debtor’s patrimonial transfers, the debtor being in insolvency procedure

    Directory of Open Access Journals (Sweden)

    Angela MIFF

    2015-06-01

    Full Text Available The study aims to present relevant aspects from the jurisprudential solutions of the High Court of Cassation and Justice in matters of the action in annulment of the patrimonial transfers of the debtor in insolvency procedure and of other judicial actions introduced by the procedure bodies or, by case, the participants to the procedure who are enabled to use the measures prescribed by the law for the purpose of restoring the debtor’s patrimony. The jurisprudential solutions adopted under the incidence of the former Law no.85/2006 on the insolvency procedure represent, in present, under diverse aspects, elements of continuity with the principles and rules established by the new law in this matter. Law no.85/2014 on the prevention insolvency procedures and of insolvency establishes the legal frame for the exertion of the measures having as purpose the restoration, in the debtor’s patrimony, of certain assets, transferred by the debtor to the fraud of the creditors’ interests, or of their value, in the scope of covering the passive part to satisfy the creditors’ interests. The new law continues the tradition of the former regulations in this matter but, also, brings some novelty elements such as the decrease or, in some cases, the increase of the duration of certain terms that the exertion of the mentioned judicial actions or their object refer to or the completion of the category of the persons entitled to introduce the mentioned judicial actions with the creditor who holds more than 50% of the value of the claims enlisted in the amount of claims.

  18. In-Orbit Vicarious Calibration for Ocean Color and Aerosol Products

    National Research Council Canada - National Science Library

    Wang, Menghua

    2005-01-01

    It is well known that, to accurately retrieve the spectrum of the water-leaving radiance and derive the ocean color products from satellite sensors, a vicarious calibration procedure, which performs...

  19. The Pelindaba facility for calibrating radiometric field instruments

    International Nuclear Information System (INIS)

    Corner, B.; Toens, P.D.; Van As, D.; Vleggaar, C.M.; Richards, D.J.

    1979-04-01

    The tremendous upsurge in uranium exploration activity, experienced in recent years, has made the need for the standardisation and calibration of radiometric field instruments apparent. In order to fulfill this need, construction of a calibration facility at the National Nuclear Research Centre, Pelindaba, was commenced in 1972 and has since been extended according the the requirements of the mining industry. The facility currently comprises 11 surface standard sources suitable for the calibration, in terms of radio-element concentration, of portable scintillometers and spectrometers, and single uranium and thorium model-borehole sources which make possible the accurate calibration of borehole logging instruments both for gross-count and spectrometric surveys. Portable potassium, uranium and thorium sources are also available for the purposes of establishing airborne-spectrometer stripping ratios. The relevant physico-chemical properties of the standards are presented in this report and calibration procedures and data reduction techniques recommended. Examples are given of in situ measurements, both on surface and down-the-hole, which show that the derived calibration constants yield radiometric grades which are, on average, accurate to within 5% of the true radio-element concentrations. A secondary facility comprising single borehole- and surface-uranium sources has also been constructed in Beaufort West in the southern Karoo [af

  20. Expert Performance Transfer - Making Knowledge Transfer Count with ExPerT Methodology

    International Nuclear Information System (INIS)

    Turner, C.L.; Braudt, T.E.

    2011-01-01

    'Knowledge Transfer' is a high-priority imperative as the nuclear industry faces the combined effects of an aging workforce and economic pressures to do more with less. Knowledge Transfer is only a part of the solution to these challenges, however. The more compelling and immediate need faced by industry is Accomplishment Transfer, or the transference of the applied knowledge necessary to assure optimal performance transfer from experienced, high-performing staff to inexperienced staff. A great deal of industry knowledge and required performance information has been documented in the form of procedures. Often under-appreciated either as knowledge stores or as drivers of human performance, procedures, coupled with tightly-focused and effective training, are arguably the most effective influences on human and plant performance. (author)

  1. Community Radiative Transfer Model for Inter-Satellites Calibration and Verification

    Science.gov (United States)

    Liu, Q.; Nalli, N. R.; Ignatov, A.; Garrett, K.; Chen, Y.; Weng, F.; Boukabara, S. A.; van Delst, P. F.; Groff, D. N.; Collard, A.; Joseph, E.; Morris, V. R.; Minnett, P. J.

    2014-12-01

    Developed at the Joint Center for Satellite Data Assimilation, the Community Radiative Transfer Model (CRTM) [1], operationally supports satellite radiance assimilation for weather forecasting. The CRTM also supports JPSS/NPP and GOES-R missions [2] for instrument calibration, validation, monitoring long-term trending, and satellite retrieved products [3]. The CRTM is used daily at the NOAA NCEP to quantify the biases and standard deviations between radiance simulations and satellite radiance measurements in a time series and angular dependency. The purposes of monitoring the data assimilation system are to ensure the proper performance of the assimilation system and to diagnose problems with the system for future improvements. The CRTM is a very useful tool for cross-sensor verifications. Using the double difference method, it can remove the biases caused by slight differences in spectral response and geometric angles between measurements of the two instruments. The CRTM is particularly useful to reduce the difference between instruments for climate studies [4]. In this study, we will carry out the assessment of the Suomi National Polar-orbiting Partnership (SNPP) [5] Cross-track Infrared Sounder (CrIS) data [6], Advanced Technology Microwave Sounder (ATMS) data, and data for Visible Infrared Imaging Radiometer Suite (VIIRS) [7][8] thermal emissive bands. We use dedicated radiosondes and surface data acquired from NOAA Aerosols and Ocean Science Expeditions (AEROSE) [9]. The high quality radiosondes were launched when Suomi NPP flew over NOAA Ship Ronald H. Brown situated in the tropical Atlantic Ocean. The atmospheric data include profiles of temperature, water vapor, and ozone, as well as total aerosol optical depths. The surface data includes air temperature and humidity at 2 meters, skin temperature (Marine Atmospheric Emitted Radiance Interferometer, M-AERI [10]), surface temperature, and surface wind vector. [1] Liu, Q., and F. Weng, 2006: JAS [2] Liu, Q

  2. Decoder calibration with ultra small current sample set for intracortical brain-machine interface

    Science.gov (United States)

    Zhang, Peng; Ma, Xuan; Chen, Luyao; Zhou, Jin; Wang, Changyong; Li, Wei; He, Jiping

    2018-04-01

    Objective. Intracortical brain-machine interfaces (iBMIs) aim to restore efficient communication and movement ability for paralyzed patients. However, frequent recalibration is required for consistency and reliability, and every recalibration will require relatively large most current sample set. The aim in this study is to develop an effective decoder calibration method that can achieve good performance while minimizing recalibration time. Approach. Two rhesus macaques implanted with intracortical microelectrode arrays were trained separately on movement and sensory paradigm. Neural signals were recorded to decode reaching positions or grasping postures. A novel principal component analysis-based domain adaptation (PDA) method was proposed to recalibrate the decoder with only ultra small current sample set by taking advantage of large historical data, and the decoding performance was compared with other three calibration methods for evaluation. Main results. The PDA method closed the gap between historical and current data effectively, and made it possible to take advantage of large historical data for decoder recalibration in current data decoding. Using only ultra small current sample set (five trials of each category), the decoder calibrated using the PDA method could achieve much better and more robust performance in all sessions than using other three calibration methods in both monkeys. Significance. (1) By this study, transfer learning theory was brought into iBMIs decoder calibration for the first time. (2) Different from most transfer learning studies, the target data in this study were ultra small sample set and were transferred to the source data. (3) By taking advantage of historical data, the PDA method was demonstrated to be effective in reducing recalibration time for both movement paradigm and sensory paradigm, indicating a viable generalization. By reducing the demand for large current training data, this new method may facilitate the application

  3. Analyses of Crawford's uvby β calibrations using the pulsational variations of FG Vir

    Science.gov (United States)

    Haas, P.

    Crawford's uvby β calibration method is examined for A-type stars by comparing it with the pulsational variations of the observable m1, c1 and MV for the δ Scuti star FG Vir. The fit between the calibration values of m1 and MV and the respective measurements for FG Vir are tested as a function of temperature based on 3068 4-colour values taken at the Observatorio de Sierra Nevada in Spain during the years 2002 and 2003. Testing is performed by means of linear regression. The fit between the measured index m1 of FG Vir and the m1 index of the Hyades is nearly perfect. A fit between the calibration value MV and the measured values of FG Vir cannot be obtained with Crawford's calibration procedure in a straightforward manner. In order to achieve an optimal fit for MV two modifications of the calibration procedure are investigated and discussed. (i) the position of the ZAMS given by Crawford is replaced by the position of the ZAMS given by Mermilliod; (ii) the influence of the mass difference on c1 is taken into account.

  4. The use of a secondary standard x-ray exposure meter to calibrate a field instrument for use in output measurements

    International Nuclear Information System (INIS)

    1974-11-01

    It is stated that the recommended procedures have been closely correlated with ICRU Report 23 (1973), 'Measurements of absorbed dose in a phantom irradiated by a single beam of X- or gamma-rays'. The present report is contained in sections entitled: introduction; equipment and outline of procedures; calibration and use of field instrument in a phantom; calibration and use of the field instrument in air; local reference instrument; summary of recommended procedures. Appendices are entitled: the Perspex Intercomparison Phantom; the implications of applying an in-air calibration factor to readings made in a water phantom. (U.K.)

  5. Optimization of SPECT calibration for quantification of images applied to dosimetry with iodine-131

    International Nuclear Information System (INIS)

    Carvalho, Samira Marques de

    2018-01-01

    SPECT systems calibration plays an essential role in the accuracy of the quantification of images. In this work, in its first stage, an optimized SPECT calibration method was proposed for 131 I studies, considering the partial volume effect (PVE) and the position of the calibration source. In the second stage, the study aimed to investigate the impact of count density and reconstruction parameters on the determination of the calibration factor and the quantification of the image in dosimetry studies, considering the reality of clinical practice in Brazil. In the final step, the study aimed evaluating the influence of several factors in the calibration for absorbed dose calculation using Monte Carlo simulations (MC) GATE code. Calibration was performed by determining a calibration curve (sensitivity versus volume) obtained by applying different thresholds. Then, the calibration factors were determined with an exponential function adjustment. Images were performed with high and low counts densities for several source positions within the simulator. To validate the calibration method, the calibration factors were used for absolute quantification of the total reference activities. The images were reconstructed adopting two approaches of different parameters, usually used in patient images. The methodology developed for the calibration of the tomographic system was easier and faster to implement than other procedures suggested to improve the accuracy of the results. The study also revealed the influence of the location of the calibration source, demonstrating better precision in the absolute quantification considering the location of the target region during the calibration of the system. The study applied in the Brazilian thyroid protocol suggests the revision of the calibration of the SPECT system, including different positions for the reference source, besides acquisitions considering the Signal to Noise Ratio (SNR) of the images. Finally, the doses obtained with the

  6. TRACEABILITY OF ON COORDINATE MEASURING MACHINES – CALIBRATION AND PERFORMANCE VERIFICATION

    DEFF Research Database (Denmark)

    De Chiffre, Leonardo; Savio, Enrico; Bariani, Paolo

    This document is used in connection with three exercises each of 45 minutes duration as a part of the course GEOMETRICAL METROLOGY AND MACHINE TESTING. The exercises concern three aspects of coordinate measurement traceability: 1) Performance verification of a CMM using a ball bar; 2) Calibration...... of an optical coordinate measuring machine; 3) Uncertainty assessment using the ISO 15530-3 “Calibrated workpieces” procedure....

  7. Two-demensional analysis of heat and mass transfer in porous media using the strongly implicit procedure

    Science.gov (United States)

    Curry, D. M.

    1974-01-01

    Numerical results of the heat and mass transfer in a porous matrix are presented. The coupled, nonlinear partial differential equations describing this physical phenomenon are solved in finite difference form for two dimensions, using a new iterative technique (the strongly implicit procedure). The influence of the external environment conditions (heating and pressure) is shown to produce two-dimensional flow in the porous matrix. Typical fluid and solid temperature distributions in the porous matrix and internal pressure distributions are presented.

  8. Determination of pesticide residue transfer rates (percent) from dried tea leaves to brewed tea.

    Science.gov (United States)

    Wang, Jian; Cheung, Wendy; Leung, Daniel

    2014-01-29

    This paper presents a study on pesticide residue transfer rates (%) from dried tea leaves to brewed tea. In the study, a brewing procedure simulated the preparation of a hot tea drink as in routine. After brewing, pesticide residues were extracted from brewed tea using a method known as QuEChERS (quick, easy, cheap, effective, rugged, and safe). An UHPLC/ESI-MS/MS method was developed and validated to identify and quantify up to 172 pesticides in both tea leaves and brewed tea samples. Quantification was achieved using matrix-matched standard calibration curves with isotopically labeled standards or a chemical analogue as internal standards, and the calibration curves consisted of six points (0.4, 2.0, 8.0, 16.0, 24.0, and 40.0 μg/L equivalent in sample). The method was validated at four concentration levels (4.0, 12, 20.0, and 32.0 μg/L equivalent in sample) using five different brewed tea matrices on two separate days per matrix. Method performance parameters included overall recovery, intermediate precision, and measurement uncertainty, which were evaluated according to a nested experimental design. Approximately, 95% of the pesticides studied had recoveries between 81 and 110%, intermediate precision ≤20%, and measurement uncertainty ≤40%. From a pilot study of 44 incurred tea samples, pesticide residues were examined for their ability to transfer from dried tea leaves to brewed tea. Each sample, both tea leaves and brewed tea, was analyzed in duplicate. Pesticides were found to have different transfer rates (%). For example, imidacloprid, methomyl, and carbendazim had transfer rates of 84.9, 83.4, and 92.4%, respectively.

  9. Update of X ray and gamma ray decay data standards for detector calibration and other applications. V. 2: Data selection, assessment and evaluation procedures

    International Nuclear Information System (INIS)

    2007-05-01

    Various factors such as source preparation and source detector geometry may affect the quality of measurements made with intrinsic germanium and other γ ray spectrometers. However, the accuracy of such measurements invariably depends on the accuracy of the efficiency versus energy calibration curve and hence on the accuracy of the decay data for the radionuclides from which calibration standard sources are prepared. Both half-lives and X and γ ray emission probabilities need to be known to good accuracy. The recommendations and report of this work are published in two volumes: Volume 1 - Recommended Decay Data, High Energy Gamma Ray Standards and Angular Correlation Coefficients; Volume 2 - Data Selection, Assessment and Evaluation Procedures. Volume 1 is a self-contained assembly of the recommended decay data covering half-lives and the X ray and γ ray emission probabilities of the selected radionuclides, and listings of various high energy γ ray standards and a set of angular correlation coefficients; these data are presented in a concise manner for rapid and easy access. More detailed technical features of the CRP are described in Volume 2, including the evaluation procedures adopted and extensive traceable explanations of the origins of the nuclear data used to produce the recommended values listed in Volume 1. This detail was judged to be essential in order to record and demonstrate the quality of the resulting data files and allow the reader to trace the origins of the nuclear data used to determine the recommended values. All evaluations were based on the available experimental data, supplemented with the judicious use of well established theory. Three types of data (half-lives, energies and emission probabilities) were compiled and evaluated (Annex II). Consideration was also given to the adoption of a number of prompt high energy γ rays from specific nuclear reactions (Annex III), as well as to using the γ-γ coincidence technique for efficiency

  10. Noninvasive method for the calibration of the peak voltage (kVp) meters

    International Nuclear Information System (INIS)

    Macedo, E.M.; Navarro, M.V.T.; Pereira, L.; Garcia, I.F.M.; Navarro, V.C.C.

    2015-01-01

    Quality control in diagnostic radiology is one of the mechanisms that minimize radiation exposure, and the measurement of tube voltage is one of the main test in these procedures. So, the calibration of non-invasive tube voltage meters is essential to maintain the metrological reliability of quality control tests. Thus, this work describes the implementation of the calibration methodology of the quantity tube peak voltage by the substitution method, using non-invasive standard meter, at LABPROSAUD-IFBA. The results showed great performance and when compared with calibrations by invasive methods, showed maximum difference of 4%, contemplated in the uncertainty ranges of the calibrations. (author)

  11. Influence of smoothing of X-ray spectra on parameters of calibration model

    International Nuclear Information System (INIS)

    Antoniak, W.; Urbanski, P.; Kowalska, E.

    1998-01-01

    Parameters of the calibration model before and after smoothing of X-ray spectra have been investigated. The calibration model was calculated using multivariate procedure - namely the partial least square regression (PLS). Investigations have been performed on an example of six sets of various standards used for calibration of some instruments based on X-ray fluorescence principle. The smoothing methods were compared: regression splines, Savitzky-Golay and Discrete Fourier Transform. The calculations were performed using a software package MATLAB and some home-made programs. (author)

  12. Calibration Tools for Measurement of Highly Enriched Uranium in Oxide and Mixed Uranium-Plutonium Oxide with a Passive-Active Neutron Drum Shuffler

    International Nuclear Information System (INIS)

    Mount, M; O'Connell, W; Cochran, C; Rinard, P

    2003-01-01

    Lawrence Livermore National Laboratory (LLNL) has completed an extensive effort to calibrate the LLNL passive-active neutron drum (PAN) shuffler (Canberra Model JCC-92) for accountability measurement of highly enriched uranium (HEU) oxide and HEU in mixed uranium-plutonium (U-Pu) oxide. Earlier papers described the PAN shuffler calibration over a range of item properties by standards measurements and an extensive series of detailed simulation calculations. With a single normalization factor, the simulations agree with the HEU oxide standards measurements to within ±1.2% at one standard deviation. Measurement errors on mixed U-Pu oxide samples are in the ±2% to ±10% range, or ±20 g for the smaller items. The purpose of this paper is to facilitate transfer of the LLNL procedure and calibration algorithms to external users who possess an identical, or equivalent, PAN shuffler. Steps include (1) measurement of HEU standards or working reference materials (WRMs); (2) MCNP simulation calculations for the standards or WRMs and a range of possible masses in the same containers; (3) a normalization of the calibration algorithms using the standard or WRM measurements to account for differences in the 252 Cf source strength, the delayed-neutron nuclear data, effects of the irradiation protocol, and detector efficiency; and (4) a verification of the simulation series trends against like LLNL results. Tools include EXCEL/Visual Basic programs which pre- and post-process the simulations, control the normalization, and embody the calibration algorithms

  13. Standard practice of calibration of force-measuring instruments for verifying the force indication of testing machines

    CERN Document Server

    American Society for Testing and Materials. Philadelphia

    2006-01-01

    1.1 The purpose of this practice is to specify procedures for the calibration of force-measuring instruments. Procedures are included for the following types of instruments: 1.1.1 Elastic force-measuring instruments, and 1.1.2 Force-multiplying systems, such as balances and small platform scales. Note 1Verification by deadweight loading is also an acceptable method of verifying the force indication of a testing machine. Tolerances for weights for this purpose are given in Practices E 4; methods for calibration of the weights are given in NIST Technical Note 577, Methods of Calibrating Weights for Piston Gages. 1.2 The values stated in SI units are to be regarded as the standard. Other metric and inch-pound values are regarded as equivalent when required. 1.3 This practice is intended for the calibration of static force measuring instruments. It is not applicable for dynamic or high speed force calibrations, nor can the results of calibrations performed in accordance with this practice be assumed valid for...

  14. Online C-arm calibration using a marked guide wire for 3D reconstruction of pulmonary arteries

    Science.gov (United States)

    Vachon, Étienne; Miró, Joaquim; Duong, Luc

    2017-03-01

    3D reconstruction of vessels from 2D X-ray angiography is highly relevant to improve the visualization and the assessment of vascular structures such as pulmonary arteries by interventional cardiologists. However, to ensure a robust and accurate reconstruction, C-arm gantry parameters must be properly calibrated to provide clinically acceptable results. Calibration procedures often rely on calibration objects and complex protocol which is not adapted to an intervention context. In this study, a novel calibration algorithm for C-arm gantry is presented using the instrumentation such as catheters and guide wire. This ensures the availability of a minimum set of correspondences and implies minimal changes to the clinical workflow. The method was evaluated on simulated data and on retrospective patient datasets. Experimental results on simulated datasets demonstrate a calibration that allows a 3D reconstruction of the guide wire up to a geometric transformation. Experiments with patients datasets show a significant decrease of the retro projection error to 0.17 mm 2D RMS. Consequently, such procedure might contribute to identify any calibration drift during the intervention.

  15. IMU-based online kinematic calibration of robot manipulator.

    Science.gov (United States)

    Du, Guanglong; Zhang, Ping

    2013-01-01

    Robot calibration is a useful diagnostic method for improving the positioning accuracy in robot production and maintenance. An online robot self-calibration method based on inertial measurement unit (IMU) is presented in this paper. The method requires that the IMU is rigidly attached to the robot manipulator, which makes it possible to obtain the orientation of the manipulator with the orientation of the IMU in real time. This paper proposed an efficient approach which incorporates Factored Quaternion Algorithm (FQA) and Kalman Filter (KF) to estimate the orientation of the IMU. Then, an Extended Kalman Filter (EKF) is used to estimate kinematic parameter errors. Using this proposed orientation estimation method will result in improved reliability and accuracy in determining the orientation of the manipulator. Compared with the existing vision-based self-calibration methods, the great advantage of this method is that it does not need the complex steps, such as camera calibration, images capture, and corner detection, which make the robot calibration procedure more autonomous in a dynamic manufacturing environment. Experimental studies on a GOOGOL GRB3016 robot show that this method has better accuracy, convenience, and effectiveness than vision-based methods.

  16. IMU-Based Online Kinematic Calibration of Robot Manipulator

    Directory of Open Access Journals (Sweden)

    Guanglong Du

    2013-01-01

    Full Text Available Robot calibration is a useful diagnostic method for improving the positioning accuracy in robot production and maintenance. An online robot self-calibration method based on inertial measurement unit (IMU is presented in this paper. The method requires that the IMU is rigidly attached to the robot manipulator, which makes it possible to obtain the orientation of the manipulator with the orientation of the IMU in real time. This paper proposed an efficient approach which incorporates Factored Quaternion Algorithm (FQA and Kalman Filter (KF to estimate the orientation of the IMU. Then, an Extended Kalman Filter (EKF is used to estimate kinematic parameter errors. Using this proposed orientation estimation method will result in improved reliability and accuracy in determining the orientation of the manipulator. Compared with the existing vision-based self-calibration methods, the great advantage of this method is that it does not need the complex steps, such as camera calibration, images capture, and corner detection, which make the robot calibration procedure more autonomous in a dynamic manufacturing environment. Experimental studies on a GOOGOL GRB3016 robot show that this method has better accuracy, convenience, and effectiveness than vision-based methods.

  17. Intensity calibrations of the broadband VUV impurity survey spectrometer - KT2

    International Nuclear Information System (INIS)

    Hawkes, N.; Peacock, N.; Lawson, K.

    1991-08-01

    Since first becoming operational in 1984 the survey spectrometer, KT2, has undergone slight modifications on the Joint European Torus Joint Undertaking (JET), and following a failure at one point the original spectrometer-'A', was exchanged for an almost identical instrument-'B'. Periodically, calibrations have been performed on the diagnostic using the diverse techniques of charge exchange branching ratios, deuterium lamp transfer irradiance standard, branching ratios from visible transitions, VUV transfer radiance standard and model calculations of line intensities in low Z ions from JET. Comparisons have been made with the theoretical instrument performance and with the prototype instruments of similar construction. This report summarises these various calibrations, carried out by the Culham Task Agreement team, until the end of 1990 when the responsability for the operation of the diagnostic was handed over to JET staff. (author)

  18. Receiver calibration and the nonlinearity parameter measurement of thick solid samples with diffraction and attenuation corrections.

    Science.gov (United States)

    Jeong, Hyunjo; Barnard, Daniel; Cho, Sungjong; Zhang, Shuzeng; Li, Xiongbing

    2017-11-01

    This paper presents analytical and experimental techniques for accurate determination of the nonlinearity parameter (β) in thick solid samples. When piezoelectric transducers are used for β measurements, the receiver calibration is required to determine the transfer function from which the absolute displacement can be calculated. The measured fundamental and second harmonic displacement amplitudes should be modified to account for beam diffraction and material absorption. All these issues are addressed in this study and the proposed technique is validated through the β measurements of thick solid samples. A simplified self-reciprocity calibration procedure for a broadband receiver is described. The diffraction and attenuation corrections for the fundamental and second harmonics are explicitly derived. Aluminum alloy samples in five different thicknesses (4, 6, 8, 10, 12cm) are prepared and β measurements are made using the finite amplitude, through-transmission method. The effects of diffraction and attenuation corrections on β measurements are systematically investigated. When diffraction and attenuation corrections are all properly made, the variation of β between different thickness samples is found to be less than 3.2%. Copyright © 2017 Elsevier B.V. All rights reserved.

  19. Novel real-time alignment and calibration of the LHCb detector in Run II

    Energy Technology Data Exchange (ETDEWEB)

    Xu, Z., E-mail: zhirui.xu@epfl.ch; Tobin, M.

    2016-07-11

    An automatic real-time alignment and calibration strategy of the LHCb detector was developed for the Run II. Thanks to the online calibration, tighter event selection criteria can be used in the trigger. Furthermore, the online calibration facilitates the use of hadronic particle identification using the Ring Imaging Cherenkov (RICH) detectors at the trigger level. The motivation for a real-time alignment and calibration of the LHCb detector is discussed from both the operational and physics performance points of view. Specific challenges of this novel configuration are discussed, as well as the working procedures of the framework and its performance.

  20. Novel real-time alignment and calibration of the LHCb detector in Run II

    CERN Document Server

    AUTHOR|(CDS)2086132; Tobin, Mark

    2016-01-01

    An automatic real-time alignment and calibration strategy of the LHCb detector was developed for the Run II. Thanks to the online calibration, tighter event selection criteria can be used in the trigger. Furthermore, the online calibration facilitates the use of hadronic particle identification using the Ring Imaging Cherenkov (RICH) detectors at the trigger level. The motivation for a real-time alignment and calibration of the LHCb detector is discussed from both the operational and physics performance points of view. Specific challenges of this novel configuration are discussed, as well as the working procedures of the framework and its performance.

  1. Calibration-measurement unit for the automation of vector network analyzer measurements

    Directory of Open Access Journals (Sweden)

    I. Rolfes

    2008-05-01

    Full Text Available With the availability of multi-port vector network analyzers, the need for automated, calibrated measurement facilities increases. In this contribution, a calibration-measurement unit is presented which realizes a repeatable automated calibration of the measurement setup as well as a user-friendly measurement of the device under test (DUT. In difference to commercially available calibration units, which are connected to the ports of the vector network analyzer preceding a measurement and which are then removed so that the DUT can be connected, the presented calibration-measurement unit is permanently connected to the ports of the VNA for the calibration as well as for the measurement of the DUT. This helps to simplify the calibrated measurement of complex scattering parameters. Moreover, a full integration of the calibration unit into the analyzer setup becomes possible. The calibration-measurement unit is based on a multiport switch setup of e.g. electromechanical relays. Under the assumption of symmetry of a switch, on the one hand the unit realizes the connection of calibration standards like one-port reflection standards and two-port through connections between different ports and on the other hand it enables the connection of the DUT. The calibration-measurement unit is applicable for two-port VNAs as well as for multiport VNAs. For the calibration of the unit, methods with completely known calibration standards like SOLT (short, open, load, through as well as self-calibration procedures like TMR or TLR can be applied.

  2. LHCb : Novel real-time alignment and calibration of the LHCb Detector in Run2

    CERN Multimedia

    Tobin, Mark

    2015-01-01

    LHCb has introduced a novel real-time detector alignment and calibration strategy for LHC Run 2. Data collected at the start of the fill will be processed in a few minutes and used to update the alignment, while the calibration constants will be evaluated for each run. This procedure will improve the quality of the online alignment. For example, the vertex locator is retracted and reinserted for stable beam collisions in each fill to be centred on the primary vertex position in the transverse plane. Consequently its position changes on a fill-by-fill basis. Critically, this new realtime alignment and calibration procedure allows identical constants to be used in the online and offline reconstruction, thus improving the correlation between triggered and offline selected events. This offers the opportunity to optimise the event selection in the trigger by applying stronger constraints. The online calibration facilitates the use of hadronic particle identification using the RICH detectors at the trigger level. T...

  3. Calibration of the CMS Drift Tube Chambers and Measurement of the Drift Velocity with Cosmic Rays

    CERN Document Server

    Chatrchyan, S; Sirunyan, A M; Adam, W; Arnold, B; Bergauer, H; Bergauer, T; Dragicevic, M; Eichberger, M; Erö, J; Friedl, M; Frühwirth, R; Ghete, V M; Hammer, J; Hänsel, S; Hoch, M; Hörmann, N; Hrubec, J; Jeitler, M; Kasieczka, G; Kastner, K; Krammer, M; Liko, D; Magrans de Abril, I; Mikulec, I; Mittermayr, F; Neuherz, B; Oberegger, M; Padrta, M; Pernicka, M; Rohringer, H; Schmid, S; Schöfbeck, R; Schreiner, T; Stark, R; Steininger, H; Strauss, J; Taurok, A; Teischinger, F; Themel, T; Uhl, D; Wagner, P; Waltenberger, W; Walzel, G; Widl, E; Wulz, C E; Chekhovsky, V; Dvornikov, O; Emeliantchik, I; Litomin, A; Makarenko, V; Marfin, I; Mossolov, V; Shumeiko, N; Solin, A; Stefanovitch, R; Suarez Gonzalez, J; Tikhonov, A; Fedorov, A; Karneyeu, A; Korzhik, M; Panov, V; Zuyeuski, R; Kuchinsky, P; Beaumont, W; Benucci, L; Cardaci, M; De Wolf, E A; Delmeire, E; Druzhkin, D; Hashemi, M; Janssen, X; Maes, T; Mucibello, L; Ochesanu, S; Rougny, R; Selvaggi, M; Van Haevermaet, H; Van Mechelen, P; Van Remortel, N; Adler, V; Beauceron, S; Blyweert, S; D'Hondt, J; De Weirdt, S; Devroede, O; Heyninck, J; Kalogeropoulos, A; Maes, J; Maes, M; Mozer, M U; Tavernier, S; Van Doninck, W; Van Mulders, P; Villella, I; Bouhali, O; Chabert, E C; Charaf, O; Clerbaux, B; De Lentdecker, G; Dero, V; Elgammal, S; Gay, A P R; Hammad, G H; Marage, P E; Rugovac, S; Vander Velde, C; Vanlaer, P; Wickens, J; Grunewald, M; Klein, B; Marinov, A; Ryckbosch, D; Thyssen, F; Tytgat, M; Vanelderen, L; Verwilligen, P; Basegmez, S; Bruno, G; Caudron, J; Delaere, C; Demin, P; Favart, D; Giammanco, A; Grégoire, G; Lemaitre, V; Militaru, O; Ovyn, S; Piotrzkowski, K; Quertenmont, L; Schul, N; Beliy, N; Daubie, E; Alves, G A; Pol, M E; Souza, M H G; Carvalho, W; De Jesus Damiao, D; De Oliveira Martins, C; Fonseca De Souza, S; Mundim, L; Oguri, V; Santoro, A; Silva Do Amaral, S M; Sznajder, A; Fernandez Perez Tomei, T R; Ferreira Dias, M A; Gregores, E M; Novaes, S F; Abadjiev, K; Anguelov, T; Damgov, J; Darmenov, N; Dimitrov, L; Genchev, V; Iaydjiev, P; Piperov, S; Stoykova, S; Sultanov, G; Trayanov, R; Vankov, I; Dimitrov, A; Dyulendarova, M; Kozhuharov, V; Litov, L; Marinova, E; Mateev, M; Pavlov, B; Petkov, P; Toteva, Z; Chen, G M; Chen, H S; Guan, W; Jiang, C H; Liang, D; Liu, B; Meng, X; Tao, J; Wang, J; Wang, Z; Xue, Z; Zhang, Z; Ban, Y; Cai, J; Ge, Y; Guo, S; Hu, Z; Mao, Y; Qian, S J; Teng, H; Zhu, B; Avila, C; Baquero Ruiz, M; Carrillo Montoya, C A; Gomez, A; Gomez Moreno, B; Ocampo Rios, A A; Osorio Oliveros, A F; Reyes Romero, D; Sanabria, J C; Godinovic, N; Lelas, K; Plestina, R; Polic, D; Puljak, I; Antunovic, Z; Dzelalija, M; Brigljevic, V; Duric, S; Kadija, K; Morovic, S; Fereos, R; Galanti, M; Mousa, J; Papadakis, A; Ptochos, F; Razis, P A; Tsiakkouri, D; Zinonos, Z; Hektor, A; Kadastik, M; Kannike, K; Müntel, M; Raidal, M; Rebane, L; Anttila, E; Czellar, S; Härkönen, J; Heikkinen, A; Karimäki, V; Kinnunen, R; Klem, J; Kortelainen, M J; Lampén, T; Lassila-Perini, K; Lehti, S; Lindén, T; Luukka, P; Mäenpää, T; Nysten, J; Tuominen, E; Tuominiemi, J; Ungaro, D; Wendland, L; Banzuzi, K; Korpela, A; Tuuva, T; Nedelec, P; Sillou, D; Besancon, M; Chipaux, R; Dejardin, M; Denegri, D; Descamps, J; Fabbro, B; Faure, J L; Ferri, F; Ganjour, S; Gentit, F X; Givernaud, A; Gras, P; Hamel de Monchenault, G; Jarry, P; Lemaire, M C; Locci, E; Malcles, J; Marionneau, M; Millischer, L; Rander, J; Rosowsky, A; Rousseau, D; Titov, M; Verrecchia, P; Baffioni, S; Bianchini, L; Bluj, M; Busson, P; Charlot, C; Dobrzynski, L; Granier de Cassagnac, R; Haguenauer, M; Miné, P; Paganini, P; Sirois, Y; Thiebaux, C; Zabi, A; Agram, J L; Besson, A; Bloch, D; Bodin, D; Brom, J M; Conte, E; Drouhin, F; Fontaine, J C; Gelé, D; Goerlach, U; Gross, L; Juillot, P; Le Bihan, A C; Patois, Y; Speck, J; Van Hove, P; Baty, C; Bedjidian, M; Blaha, J; Boudoul, G; Brun, H; Chanon, N; Chierici, R; Contardo, D; Depasse, P; Dupasquier, T; El Mamouni, H; Fassi, F; Fay, J; Gascon, S; Ille, B; Kurca, T; Le Grand, T; Lethuillier, M; Lumb, N; Mirabito, L; Perries, S; Vander Donckt, M; Verdier, P; Djaoshvili, N; Roinishvili, N; Roinishvili, V; Amaglobeli, N; Adolphi, R; Anagnostou, G; Brauer, R; Braunschweig, W; Edelhoff, M; Esser, H; Feld, L; Karpinski, W; Khomich, A; Klein, K; Mohr, N; Ostaptchouk, A; Pandoulas, D; Pierschel, G; Raupach, F; Schael, S; Schultz von Dratzig, A; Schwering, G; Sprenger, D; Thomas, M; Weber, M; Wittmer, B; Wlochal, M; Actis, O; Altenhöfer, G; Bender, W; Biallass, P; Erdmann, M; Fetchenhauer, G; Frangenheim, J; Hebbeker, T; Hilgers, G; Hinzmann, A; Hoepfner, K; Hof, C; Kirsch, M; Klimkovich, T; Kreuzer, P; Lanske, D; Merschmeyer, M; Meyer, A; Philipps, B; Pieta, H; Reithler, H; Schmitz, S A; Sonnenschein, L; Sowa, M; Steggemann, J; Szczesny, H; Teyssier, D; Zeidler, C; Bontenackels, M; Davids, M; Duda, M; Flügge, G; Geenen, H; Giffels, M; Haj Ahmad, W; Hermanns, T; Heydhausen, D; Kalinin, S; Kress, T; Linn, A; Nowack, A; Perchalla, L; Poettgens, M; Pooth, O; Sauerland, P; Stahl, A; Tornier, D; Zoeller, M H; Aldaya Martin, M; Behrens, U; Borras, K; Campbell, A; Castro, E; Dammann, D; Eckerlin, G; Flossdorf, A; Flucke, G; Geiser, A; Hatton, D; Hauk, J; Jung, H; Kasemann, M; Katkov, I; Kleinwort, C; Kluge, H; Knutsson, A; Kuznetsova, E; Lange, W; Lohmann, W; Mankel, R; Marienfeld, M; Meyer, A B; Miglioranzi, S; Mnich, J; Ohlerich, M; Olzem, J; Parenti, A; Rosemann, C; Schmidt, R; Schoerner-Sadenius, T; Volyanskyy, D; Wissing, C; Zeuner, W D; Autermann, C; Bechtel, F; Draeger, J; Eckstein, D; Gebbert, U; Kaschube, K; Kaussen, G; Klanner, R; Mura, B; Naumann-Emme, S; Nowak, F; Pein, U; Sander, C; Schleper, P; Schum, T; Stadie, H; Steinbrück, G; Thomsen, J; Wolf, R; Bauer, J; Blüm, P; Buege, V; Cakir, A; Chwalek, T; De Boer, W; Dierlamm, A; Dirkes, G; Feindt, M; Felzmann, U; Frey, M; Furgeri, A; Gruschke, J; Hackstein, C; Hartmann, F; Heier, S; Heinrich, M; Held, H; Hirschbuehl, D; Hoffmann, K H; Honc, S; Jung, C; Kuhr, T; Liamsuwan, T; Martschei, D; Mueller, S; Müller, Th; Neuland, M B; Niegel, M; Oberst, O; Oehler, A; Ott, J; Peiffer, T; Piparo, D; Quast, G; Rabbertz, K; Ratnikov, F; Ratnikova, N; Renz, M; Saout, C; Sartisohn, G; Scheurer, A; Schieferdecker, P; Schilling, F P; Schott, G; Simonis, H J; Stober, F M; Sturm, P; Troendle, D; Trunov, A; Wagner, W; Wagner-Kuhr, J; Zeise, M; Zhukov, V; Ziebarth, E B; Daskalakis, G; Geralis, T; Karafasoulis, K; Kyriakis, A; Loukas, D; Markou, A; Markou, C; Mavrommatis, C; Petrakou, E; Zachariadou, A; Gouskos, L; Katsas, P; Panagiotou, A; Evangelou, I; Kokkas, P; Manthos, N; Papadopoulos, I; Patras, V; Triantis, F A; Bencze, G; Boldizsar, L; Debreczeni, G; Hajdu, C; Hernath, S; Hidas, P; Horvath, D; Krajczar, K; Laszlo, A; Patay, G; Sikler, F; Toth, N; Vesztergombi, G; Beni, N; Christian, G; Imrek, J; Molnar, J; Novak, D; Palinkas, J; Szekely, G; Szillasi, Z; Tokesi, K; Veszpremi, V; Kapusi, A; Marian, G; Raics, P; Szabo, Z; Trocsanyi, Z L; Ujvari, B; Zilizi, G; Bansal, S; Bawa, H S; Beri, S B; Bhatnagar, V; Jindal, M; Kaur, M; Kaur, R; Kohli, J M; Mehta, M Z; Nishu, N; Saini, L K; Sharma, A; Singh, A; Singh, J B; Singh, S P; Ahuja, S; Arora, S; Bhattacharya, S; Chauhan, S; Choudhary, B C; Gupta, P; Jain, S; Jha, M; Kumar, A; Ranjan, K; Shivpuri, R K; Srivastava, A K; Choudhury, R K; Dutta, D; Kailas, S; Kataria, S K; Mohanty, A K; Pant, L M; Shukla, P; Topkar, A; Aziz, T; Guchait, M; Gurtu, A; Maity, M; Majumder, D; Majumder, G; Mazumdar, K; Nayak, A; Saha, A; Sudhakar, K; Banerjee, S; Dugad, S; Mondal, N K; Arfaei, H; Bakhshiansohi, H; Fahim, A; Jafari, A; Mohammadi Najafabadi, M; Moshaii, A; Paktinat Mehdiabadi, S; Rouhani, S; Safarzadeh, B; Zeinali, M; Felcini, M; Abbrescia, M; Barbone, L; Chiumarulo, F; Clemente, A; Colaleo, A; Creanza, D; Cuscela, G; De Filippis, N; De Palma, M; De Robertis, G; Donvito, G; Fedele, F; Fiore, L; Franco, M; Iaselli, G; Lacalamita, N; Loddo, F; Lusito, L; Maggi, G; Maggi, M; Manna, N; Marangelli, B; My, S; Natali, S; Nuzzo, S; Papagni, G; Piccolomo, S; Pierro, G A; Pinto, C; Pompili, A; Pugliese, G; Rajan, R; Ranieri, A; Romano, F; Roselli, G; Selvaggi, G; Shinde, Y; Silvestris, L; Tupputi, S; Zito, G; Abbiendi, G; Bacchi, W; Benvenuti, A C; Boldini, M; Bonacorsi, D; Braibant-Giacomelli, S; Cafaro, V D; Caiazza, S S; Capiluppi, P; Castro, A; Cavallo, F R; Codispoti, G; Cuffiani, M; D'Antone, I; Dallavalle, G M; Fabbri, F; Fanfani, A; Fasanella, D; Giacomelli, P; Giordano, V; Giunta, M; Grandi, C; Guerzoni, M; Marcellini, S; Masetti, G; Montanari, A; Navarria, F L; Odorici, F; Pellegrini, G; Perrotta, A; Rossi, A M; Rovelli, T; Siroli, G; Torromeo, G; Travaglini, R; Albergo, S; Costa, S; Potenza, R; Tricomi, A; Tuve, C; Barbagli, G; Broccolo, G; Ciulli, V; Civinini, C; D'Alessandro, R; Focardi, E; Frosali, S; Gallo, E; Genta, C; Landi, G; Lenzi, P; Meschini, M; Paoletti, S; Sguazzoni, G; Tropiano, A; Benussi, L; Bertani, M; Bianco, S; Colafranceschi, S; Colonna, D; Fabbri, F; Giardoni, M; Passamonti, L; Piccolo, D; Pierluigi, D; Ponzio, B; Russo, A; Fabbricatore, P; Musenich, R; Benaglia, A; Calloni, M; Cerati, G B; D'Angelo, P; De Guio, F; Farina, F M; Ghezzi, A; Govoni, P; Malberti, M; Malvezzi, S; Martelli, A; Menasce, D; Miccio, V; Moroni, L; Negri, P; Paganoni, M; Pedrini, D; Pullia, A; Ragazzi, S; Redaelli, N; Sala, S; Salerno, R; Tabarelli de Fatis, T; Tancini, V; Taroni, S; Buontempo, S; Cavallo, N; Cimmino, A; De Gruttola, M; Fabozzi, F; Iorio, A O M; Lista, L; Lomidze, D; Noli, P; Paolucci, P; Sciacca, C; Azzi, P; Bacchetta, N; Barcellan, L; Bellan, P; Bellato, M; Benettoni, M; Biasotto, M; Bisello, D; Borsato, E; Branca, A; Carlin, R; Castellani, L; Checchia, P; Conti, E; Dal Corso, F; De Mattia, M; Dorigo, T; Dosselli, U; Fanzago, F; Gasparini, F; Gasparini, U; Giubilato, P; Gonella, F; Gresele, A; Gulmini, M; Kaminskiy, A; Lacaprara, S; Lazzizzera, I; Margoni, M; Maron, G; Mattiazzo, S; Mazzucato, M; Meneghelli, M; Meneguzzo, A T; Michelotto, M; Montecassiano, F; Nespolo, M; Passaseo, M; Pegoraro, M; Perrozzi, L; Pozzobon, N; Ronchese, P; Simonetto, F; Toniolo, N; Torassa, E; Tosi, M; Triossi, A; Vanini, S; Ventura, S; Zotto, P; Zumerle, G; Baesso, P; Berzano, U; Bricola, S; Necchi, M M; Pagano, D; Ratti, S P; Riccardi, C; Torre, P; Vicini, A; Vitulo, P; Viviani, C; Aisa, D; Aisa, S; Babucci, E; Biasini, M; Bilei, G M; Caponeri, B; Checcucci, B; Dinu, N; Fanò, L; Farnesini, L; Lariccia, P; Lucaroni, A; Mantovani, G; Nappi, A; Piluso, A; Postolache, V; Santocchia, A; Servoli, L; Tonoiu, D; Vedaee, A; Volpe, R; Azzurri, P; Bagliesi, G; Bernardini, J; Berretta, L; Boccali, T; Bocci, A; Borrello, L; Bosi, F; Calzolari, F; Castaldi, R; Dell'Orso, R; Fiori, F; Foà, L; Gennai, S; Giassi, A; Kraan, A; Ligabue, F; Lomtadze, T; Mariani, F; Martini, L; Massa, M; Messineo, A; Moggi, A; Palla, F; Palmonari, F; Petragnani, G; Petrucciani, G; Raffaelli, F; Sarkar, S; Segneri, G; Serban, A T; Spagnolo, P; Tenchini, R; Tolaini, S; Tonelli, G; Venturi, A; Verdini, P G; Baccaro, S; Barone, L; Bartoloni, A; Cavallari, F; Dafinei, I; Del Re, D; Di Marco, E; Diemoz, M; Franci, D; Longo, E; Organtini, G; Palma, A; Pandolfi, F; Paramatti, R; Pellegrino, F; Rahatlou, S; Rovelli, C; Alampi, G; Amapane, N; Arcidiacono, R; Argiro, S; Arneodo, M; Biino, C; Borgia, M A; Botta, C; Cartiglia, N; Castello, R; Cerminara, G; Costa, M; Dattola, D; Dellacasa, G; Demaria, N; Dughera, G; Dumitrache, F; Graziano, A; Mariotti, C; Marone, M; Maselli, S; Migliore, E; Mila, G; Monaco, V; Musich, M; Nervo, M; Obertino, M M; Oggero, S; Panero, R; Pastrone, N; Pelliccioni, M; Romero, A; Ruspa, M; Sacchi, R; Solano, A; Staiano, A; Trapani, P P; Trocino, D; Vilela Pereira, A; Visca, L; Zampieri, A; Ambroglini, F; Belforte, S; Cossutti, F; Della Ricca, G; Gobbo, B; Penzo, A; Chang, S; Chung, J; Kim, D H; Kim, G N; Kong, D J; Park, H; Son, D C; Bahk, S Y; Song, S; Jung, S Y; Hong, B; Kim, H; Kim, J H; Lee, K S; Moon, D H; Park, S K; Rhee, H B; Sim, K S; Kim, J; Choi, M; Hahn, G; Park, I C; Choi, S; Choi, Y; Goh, J; Jeong, H; Kim, T J; Lee, J; Lee, S; Janulis, M; Martisiute, D; Petrov, P; Sabonis, T; Castilla Valdez, H; Sánchez Hernández, A; Carrillo Moreno, S; Morelos Pineda, A; Allfrey, P; Gray, R N C; Krofcheck, D; Bernardino Rodrigues, N; Butler, P H; Signal, T; Williams, J C; Ahmad, M; Ahmed, I; Ahmed, W; Asghar, M I; Awan, M I M; Hoorani, H R; Hussain, I; Khan, W A; Khurshid, T; Muhammad, S; Qazi, S; Shahzad, H; Cwiok, M; Dabrowski, R; Dominik, W; Doroba, K; Konecki, M; Krolikowski, J; Pozniak, K; Romaniuk, Ryszard; Zabolotny, W; Zych, P; Frueboes, T; Gokieli, R; Goscilo, L; Górski, M; Kazana, M; Nawrocki, K; Szleper, M; Wrochna, G; Zalewski, P; Almeida, N; Antunes Pedro, L; Bargassa, P; David, A; Faccioli, P; Ferreira Parracho, P G; Freitas Ferreira, M; Gallinaro, M; Guerra Jordao, M; Martins, P; Mini, G; Musella, P; Pela, J; Raposo, L; Ribeiro, P Q; Sampaio, S; Seixas, J; Silva, J; Silva, P; Soares, D; Sousa, M; Varela, J; Wöhri, H K; Altsybeev, I; Belotelov, I; Bunin, P; Ershov, Y; Filozova, I; Finger, M; Finger, M., Jr.; Golunov, A; Golutvin, I; Gorbounov, N; Kalagin, V; Kamenev, A; Karjavin, V; Konoplyanikov, V; Korenkov, V; Kozlov, G; Kurenkov, A; Lanev, A; Makankin, A; Mitsyn, V V; Moisenz, P; Nikonov, E; Oleynik, D; Palichik, V; Perelygin, V; Petrosyan, A; Semenov, R; Shmatov, S; Smirnov, V; Smolin, D; Tikhonenko, E; Vasil'ev, S; Vishnevskiy, A; Volodko, A; Zarubin, A; Zhiltsov, V; Bondar, N; Chtchipounov, L; Denisov, A; Gavrikov, Y; Gavrilov, G; Golovtsov, V; Ivanov, Y; Kim, V; Kozlov, V; Levchenko, P; Obrant, G; Orishchin, E; Petrunin, A; Shcheglov, Y; Shchetkovskiy, A; Sknar, V; Smirnov, I; Sulimov, V; Tarakanov, V; Uvarov, L; Vavilov, S; Velichko, G; Volkov, S; Vorobyev, A; Andreev, Yu; Anisimov, A; Antipov, P; Dermenev, A; Gninenko, S; Golubev, N; Kirsanov, M; Krasnikov, N; Matveev, V; Pashenkov, A; Postoev, V E; Solovey, A; Toropin, A; Troitsky, S; Baud, A; Epshteyn, V; Gavrilov, V; Ilina, N; Kaftanov, V; Kolosov, V; Kossov, M; Krokhotin, A; Kuleshov, S; Oulianov, A; Safronov, G; Semenov, S; Shreyber, I; Stolin, V; Vlasov, E; Zhokin, A; Boos, E; Dubinin, M; Dudko, L; Ershov, A; Gribushin, A; Klyukhin, V; Kodolova, O; Lokhtin, I; Petrushanko, S; Sarycheva, L; Savrin, V; Snigirev, A; Vardanyan, I; Dremin, I; Kirakosyan, M; Konovalova, N; Rusakov, S V; Vinogradov, A; Akimenko, S; Artamonov, A; Azhgirey, I; Bitioukov, S; Burtovoy, V; Grishin, V; Kachanov, V; Konstantinov, D; Krychkine, V; Levine, A; Lobov, I; Lukanin, V; Mel'nik, Y; Petrov, V; Ryutin, R; Slabospitsky, S; Sobol, A; Sytine, A; Tourtchanovitch, L; Troshin, S; Tyurin, N; Uzunian, A; Volkov, A; Adzic, P; Djordjevic, M; Jovanovic, D; Krpic, D; Maletic, D; Puzovic, J; Smiljkovic, N; Aguilar-Benitez, M; Alberdi, J; Alcaraz Maestre, J; Arce, P; Barcala, J M; Battilana, C; Burgos Lazaro, C; Caballero Bejar, J; Calvo, E; Cardenas Montes, M; Cepeda, M; Cerrada, M; Chamizo Llatas, M; Clemente, F; Colino, N; Daniel, M; De La Cruz, B; Delgado Peris, A; Diez Pardos, C; Fernandez Bedoya, C; Fernández Ramos, J P; Ferrando, A; Flix, J; Fouz, M C; Garcia-Abia, P; Garcia-Bonilla, A C; Gonzalez Lopez, O; Goy Lopez, S; Hernandez, J M; Josa, M I; Marin, J; Merino, G; Molina, J; Molinero, A; Navarrete, J J; Oller, J C; Puerta Pelayo, J; Romero, L; Santaolalla, J; Villanueva Munoz, C; Willmott, C; Yuste, C; Albajar, C; Blanco Otano, M; de Trocóniz, J F; Garcia Raboso, A; Lopez Berengueres, J O; Cuevas, J; Fernandez Menendez, J; Gonzalez Caballero, I; Lloret Iglesias, L; Naves Sordo, H; Vizan Garcia, J M; Cabrillo, I J; Calderon, A; Chuang, S H; Diaz Merino, I; Diez Gonzalez, C; Duarte Campderros, J; Fernandez, M; Gomez, G; Gonzalez Sanchez, J; Gonzalez Suarez, R; Jorda, C; Lobelle Pardo, P; Lopez Virto, A; Marco, J; Marco, R; Martinez Rivero, C; Martinez Ruiz del Arbol, P; Matorras, F; Rodrigo, T; Ruiz Jimeno, A; Scodellaro, L; Sobron Sanudo, M; Vila, I; Vilar Cortabitarte, R; Abbaneo, D; Albert, E; Alidra, M; Ashby, S; Auffray, E; Baechler, J; Baillon, P; Ball, A H; Bally, S L; Barney, D; Beaudette, F; Bellan, R; Benedetti, D; Benelli, G; Bernet, C; Bloch, P; Bolognesi, S; Bona, M; Bos, J; Bourgeois, N; Bourrel, T; Breuker, H; Bunkowski, K; Campi, D; Camporesi, T; Cano, E; Cattai, A; Chatelain, J P; Chauvey, M; Christiansen, T; Coarasa Perez, J A; Conde Garcia, A; Covarelli, R; Curé, B; De Roeck, A; Delachenal, V; Deyrail, D; Di Vincenzo, S; Dos Santos, S; Dupont, T; Edera, L M; Elliott-Peisert, A; Eppard, M; Favre, M; Frank, N; Funk, W; Gaddi, A; Gastal, M; Gateau, M; Gerwig, H; Gigi, D; Gill, K; Giordano, D; Girod, J P; Glege, F; Gomez-Reino Garrido, R; Goudard, R; Gowdy, S; Guida, R; Guiducci, L; Gutleber, J; Hansen, M; Hartl, C; Harvey, J; Hegner, B; Hoffmann, H F; Holzner, A; Honma, A; Huhtinen, M; Innocente, V; Janot, P; Le Godec, G; Lecoq, P; Leonidopoulos, C; Loos, R; Lourenço, C; Lyonnet, A; Macpherson, A; Magini, N; Maillefaud, J D; Maire, G; Mäki, T; Malgeri, L; Mannelli, M; Masetti, L; Meijers, F; Meridiani, P; Mersi, S; Meschi, E; Meynet Cordonnier, A; Moser, R; Mulders, M; Mulon, J; Noy, M; Oh, A; Olesen, G; Onnela, A; Orimoto, T; Orsini, L; Perez, E; Perinic, G; Pernot, J F; Petagna, P; Petiot, P; Petrilli, A; Pfeiffer, A; Pierini, M; Pimiä, M; Pintus, R; Pirollet, B; Postema, H; Racz, A; Ravat, S; Rew, S B; Rodrigues Antunes, J; Rolandi, G; Rovere, M; Ryjov, V; Sakulin, H; Samyn, D; Sauce, H; Schäfer, C; Schlatter, W D; Schröder, M; Schwick, C; Sciaba, A; Segoni, I; Sharma, A; Siegrist, N; Siegrist, P; Sinanis, N; Sobrier, T; Sphicas, P; Spiga, D; Spiropulu, M; Stöckli, F; Traczyk, P; Tropea, P; Troska, J; Tsirou, A; Veillet, L; Veres, G I; Voutilainen, M; Wertelaers, P; Zanetti, M; Bertl, W; Deiters, K; Erdmann, W; Gabathuler, K; Horisberger, R; Ingram, Q; Kaestli, H C; König, S; Kotlinski, D; Langenegger, U; Meier, F; Renker, D; Rohe, T; Sibille, J; Starodumov, A; Betev, B; Caminada, L; Chen, Z; Cittolin, S; Da Silva Di Calafiori, D R; Dambach, S; Dissertori, G; Dittmar, M; Eggel, C; Eugster, J; Faber, G; Freudenreich, K; Grab, C; Hervé, A; Hintz, W; Lecomte, P; Luckey, P D; Lustermann, W; Marchica, C; Milenovic, P; Moortgat, F; Nardulli, A; Nessi-Tedaldi, F; Pape, L; Pauss, F; Punz, T; Rizzi, A; Ronga, F J; Sala, L; Sanchez, A K; Sawley, M C; Sordini, V; Stieger, B; Tauscher, L; Thea, A; Theofilatos, K; Treille, D; Trüb, P; Weber, M; Wehrli, L; Weng, J; Zelepoukine, S; Amsler, C; Chiochia, V; De Visscher, S; Regenfus, C; Robmann, P; Rommerskirchen, T; Schmidt, A; Tsirigkas, D; Wilke, L; Chang, Y H; Chen, E A; Chen, W T; Go, A; Kuo, C M; Li, S W; Lin, W; Bartalini, P; Chang, P; Chao, Y; Chen, K F; Hou, W S; Hsiung, Y; Lei, Y J; Lin, S W; Lu, R S; Schümann, J; Shiu, J G; Tzeng, Y M; Ueno, K; Velikzhanin, Y; Wang, C C; Wang, M; Adiguzel, A; Ayhan, A; Azman Gokce, A; Bakirci, M N; Cerci, S; Dumanoglu, I; Eskut, E; Girgis, S; Gurpinar, E; Hos, I; Karaman, T; Kayis Topaksu, A; Kurt, P; Önengüt, G; Önengüt Gökbulut, G; Ozdemir, K; Ozturk, S; Polatöz, A; Sogut, K; Tali, B; Topakli, H; Uzun, D; Vergili, L N; Vergili, M; Akin, I V; Aliev, T; Bilmis, S; Deniz, M; Gamsizkan, H; Guler, A M; Öcalan, K; Serin, M; Sever, R; Surat, U E; Zeyrek, M; Deliomeroglu, M; Demir, D; Gülmez, E; Halu, A; Isildak, B; Kaya, M; Kaya, O; Ozkorucuklu, S; Sonmez, N; Levchuk, L; Lukyanenko, S; Soroka, D; Zub, S; Bostock, F; Brooke, J J; Cheng, T L; Cussans, D; Frazier, R; Goldstein, J; Grant, N; Hansen, M; Heath, G P; Heath, H F; Hill, C; Huckvale, B; Jackson, J; Mackay, C K; Metson, S; Newbold, D M; Nirunpong, K; Smith, V J; Velthuis, J; Walton, R; Bell, K W; Brew, C; Brown, R M; Camanzi, B; Cockerill, D J A; Coughlan, J A; Geddes, N I; Harder, K; Harper, S; Kennedy, B W; Murray, P; Shepherd-Themistocleous, C H; Tomalin, I R; Williams, J H; Womersley, W J; Worm, S D; Bainbridge, R; Ball, G; Ballin, J; Beuselinck, R; Buchmuller, O; Colling, D; Cripps, N; Davies, G; Della Negra, M; Foudas, C; Fulcher, J; Futyan, D; Hall, G; Hays, J; Iles, G; Karapostoli, G; MacEvoy, B C; Magnan, A M; Marrouche, J; Nash, J; Nikitenko, A; Papageorgiou, A; Pesaresi, M; Petridis, K; Pioppi, M; Raymond, D M; Rompotis, N; Rose, A; Ryan, M J; Seez, C; Sharp, P; Sidiropoulos, G; Stettler, M; Stoye, M; Takahashi, M; Tapper, A; Timlin, C; Tourneur, S; Vazquez Acosta, M; Virdee, T; Wakefield, S; Wardrope, D; Whyntie, T; Wingham, M; Cole, J E; Goitom, I; Hobson, P R; Khan, A; Kyberd, P; Leslie, D; Munro, C; Reid, I D; Siamitros, C; Taylor, R; Teodorescu, L; Yaselli, I; Bose, T; Carleton, M; Hazen, E; Heering, A H; Heister, A; John, J St; Lawson, P; Lazic, D; Osborne, D; Rohlf, J; Sulak, L; Wu, S; Andrea, J; Avetisyan, A; Bhattacharya, S; Chou, J P; Cutts, D; Esen, S; Kukartsev, G; Landsberg, G; Narain, M; Nguyen, D; Speer, T; Tsang, K V; Breedon, R; Calderon De La Barca Sanchez, M; Case, M; Cebra, D; Chertok, M; Conway, J; Cox, P T; Dolen, J; Erbacher, R; Friis, E; Ko, W; Kopecky, A; Lander, R; Lister, A; Liu, H; Maruyama, S; Miceli, T; Nikolic, M; Pellett, D; Robles, J; Searle, M; Smith, J; Squires, M; Stilley, J; Tripathi, M; Vasquez Sierra, R; Veelken, C; Andreev, V; Arisaka, K; Cline, D; Cousins, R; Erhan, S; Hauser, J; Ignatenko, M; Jarvis, C; Mumford, J; Plager, C; Rakness, G; Schlein, P; Tucker, J; Valuev, V; Wallny, R; Yang, X; Babb, J; Bose, M; Chandra, A; Clare, R; Ellison, J A; Gary, J W; Hanson, G; Jeng, G Y; Kao, S C; Liu, F; Liu, H; Luthra, A; Nguyen, H; Pasztor, G; Satpathy, A; Shen, B C; Stringer, R; Sturdy, J; Sytnik, V; Wilken, R; Wimpenny, S; Branson, J G; Dusinberre, E; Evans, D; Golf, F; Kelley, R; Lebourgeois, M; Letts, J; Lipeles, E; Mangano, B; Muelmenstaedt, J; Norman, M; Padhi, S; Petrucci, A; Pi, H; Pieri, M; Ranieri, R; Sani, M; Sharma, V; Simon, S; Würthwein, F; Yagil, A; Campagnari, C; D'Alfonso, M; Danielson, T; Garberson, J; Incandela, J; Justus, C; Kalavase, P; Koay, S A; Kovalskyi, D; Krutelyov, V; Lamb, J; Lowette, S; Pavlunin, V; Rebassoo, F; Ribnik, J; Richman, J; Rossin, R; Stuart, D; To, W; Vlimant, J R; Witherell, M; Apresyan, A; Bornheim, A; Bunn, J; Chiorboli, M; Gataullin, M; Kcira, D; Litvine, V; Ma, Y; Newman, H B; Rogan, C; Timciuc, V; Veverka, J; Wilkinson, R; Yang, Y; Zhang, L; Zhu, K; Zhu, R Y; Akgun, B; Carroll, R; Ferguson, T; Jang, D W; Jun, S Y; Paulini, M; Russ, J; Terentyev, N; Vogel, H; Vorobiev, I; Cumalat, J P; Dinardo, M E; Drell, B R; Ford, W T; Heyburn, B; Luiggi Lopez, E; Nauenberg, U; Stenson, K; Ulmer, K; Wagner, S R; Zang, S L; Agostino, L; Alexander, J; Blekman, F; Cassel, D; Chatterjee, A; Das, S; Gibbons, L K; Heltsley, B; Hopkins, W; Khukhunaishvili, A; Kreis, B; Kuznetsov, V; Patterson, J R; Puigh, D; Ryd, A; Shi, X; Stroiney, S; Sun, W; Teo, W D; Thom, J; Vaughan, J; Weng, Y; Wittich, P; Beetz, C P; Cirino, G; Sanzeni, C; Winn, D; Abdullin, S; Afaq, M A; Albrow, M; Ananthan, B; Apollinari, G; Atac, M; Badgett, W; Bagby, L; Bakken, J A; Baldin, B; Banerjee, S; Banicz, K; Bauerdick, L A T; Beretvas, A; Berryhill, J; Bhat, P C; Biery, K; Binkley, M; Bloch, I; Borcherding, F; Brett, A M; Burkett, K; Butler, J N; Chetluru, V; Cheung, H W K; Chlebana, F; Churin, I; Cihangir, S; Crawford, M; Dagenhart, W; Demarteau, M; Derylo, G; Dykstra, D; Eartly, D P; Elias, J E; Elvira, V D; Evans, D; Feng, L; Fischler, M; Fisk, I; Foulkes, S; Freeman, J; Gartung, P; Gottschalk, E; Grassi, T; Green, D; Guo, Y; Gutsche, O; Hahn, A; Hanlon, J; Harris, R M; Holzman, B; Howell, J; Hufnagel, D; James, E; Jensen, H; Johnson, M; Jones, C D; Joshi, U; Juska, E; Kaiser, J; Klima, B; Kossiakov, S; Kousouris, K; Kwan, S; Lei, C M; Limon, P; Lopez Perez, J A; Los, S; Lueking, L; Lukhanin, G; Lusin, S; Lykken, J; Maeshima, K; Marraffino, J M; Mason, D; McBride, P; Miao, T; Mishra, K; Moccia, S; Mommsen, R; Mrenna, S; Muhammad, A S; Newman-Holmes, C; Noeding, C; O'Dell, V; Prokofyev, O; Rivera, R; Rivetta, C H; Ronzhin, A; Rossman, P; Ryu, S; Sekhri, V; Sexton-Kennedy, E; Sfiligoi, I; Sharma, S; Shaw, T M; Shpakov, D; Skup, E; Smith, R P; Soha, A; Spalding, W J; Spiegel, L; Suzuki, I; Tan, P; Tanenbaum, W; Tkaczyk, S; Trentadue, R; Uplegger, L; Vaandering, E W; Vidal, R; Whitmore, J; Wicklund, E; Wu, W; Yarba, J; Yumiceva, F; Yun, J C; Acosta, D; Avery, P; Barashko, V; Bourilkov, D; Chen, M; Di Giovanni, G P; Dobur, D; Drozdetskiy, A; Field, R D; Fu, Y; Furic, I K; Gartner, J; Holmes, D; Kim, B; Klimenko, S; Konigsberg, J; Korytov, A; Kotov, K; Kropivnitskaya, A; Kypreos, T; Madorsky, A; Matchev, K; Mitselmakher, G; Pakhotin, Y; Piedra Gomez, J; Prescott, C; Rapsevicius, V; Remington, R; Schmitt, M; Scurlock, B; Wang, D; Yelton, J; Ceron, C; Gaultney, V; Kramer, L; Lebolo, L M; Linn, S; Markowitz, P; Martinez, G; Rodriguez, J L; Adams, T; Askew, A; Baer, H; Bertoldi, M; Chen, J; Dharmaratna, W G D; Gleyzer, S V; Haas, J; Hagopian, S; Hagopian, V; Jenkins, M; Johnson, K F; Prettner, E; Prosper, H; Sekmen, S; Baarmand, M M; Guragain, S; Hohlmann, M; Kalakhety, H; Mermerkaya, H; Ralich, R; Vodopiyanov, I; Abelev, B; Adams, M R; Anghel, I M; Apanasevich, L; Bazterra, V E; Betts, R R; Callner, J; Castro, M A; Cavanaugh, R; Dragoiu, C; Garcia-Solis, E J; Gerber, C E; Hofman, D J; Khalatian, S; Mironov, C; Shabalina, E; Smoron, A; Varelas, N; Akgun, U; Albayrak, E A; Ayan, A S; Bilki, B; Briggs, R; Cankocak, K; Chung, K; Clarida, W; Debbins, P; Duru, F; Ingram, F D; Lae, C K; McCliment, E; Merlo, J P; Mestvirishvili, A; Miller, M J; Moeller, A; Nachtman, J; Newsom, C R; Norbeck, E; Olson, J; Onel, Y; Ozok, F; Parsons, J; Schmidt, I; Sen, S; Wetzel, J; Yetkin, T; Yi, K; Barnett, B A; Blumenfeld, B; Bonato, A; Chien, C Y; Fehling, D; Giurgiu, G; Gritsan, A V; Guo, Z J; Maksimovic, P; Rappoccio, S; Swartz, M; Tran, N V; Zhang, Y; Baringer, P; Bean, A; Grachov, O; Murray, M; Radicci, V; Sanders, S; Wood, J S; Zhukova, V; Bandurin, D; Bolton, T; Kaadze, K; Liu, A; Maravin, Y; Onoprienko, D; Svintradze, I; Wan, Z; Gronberg, J; Hollar, J; Lange, D; Wright, D; Baden, D; Bard, R; Boutemeur, M; Eno, S C; Ferencek, D; Hadley, N J; Kellogg, R G; Kirn, M; Kunori, S; Rossato, K; Rumerio, P; Santanastasio, F; Skuja, A; Temple, J; Tonjes, M B; Tonwar, S C; Toole, T; Twedt, E; Alver, B; Bauer, G; Bendavid, J; Busza, W; Butz, E; Cali, I A; Chan, M; D'Enterria, D; Everaerts, P; Gomez Ceballos, G; Hahn, K A; Harris, P; Jaditz, S; Kim, Y; Klute, M; Lee, Y J; Li, W; Loizides, C; Ma, T; Miller, M; Nahn, S; Paus, C; Roland, C; Roland, G; Rudolph, M; Stephans, G; Sumorok, K; Sung, K; Vaurynovich, S; Wenger, E A; Wyslouch, B; Xie, S; Yilmaz, Y; Yoon, A S; Bailleux, D; Cooper, S I; Cushman, P; Dahmes, B; De Benedetti, A; Dolgopolov, A; Dudero, P R; Egeland, R; Franzoni, G; Haupt, J; Inyakin, A; Klapoetke, K; Kubota, Y; Mans, J; Mirman, N; Petyt, D; Rekovic, V; Rusack, R; Schroeder, M; Singovsky, A; Zhang, J; Cremaldi, L M; Godang, R; Kroeger, R; Perera, L; Rahmat, R; Sanders, D A; Sonnek, P; Summers, D; Bloom, K; Bockelman, B; Bose, S; Butt, J; Claes, D R; Dominguez, A; Eads, M; Keller, J; Kelly, T; Kravchenko, I; Lazo-Flores, J; Lundstedt, C; Malbouisson, H; Malik, S; Snow, G R; Baur, U; Iashvili, I; Kharchilava, A; Kumar, A; Smith, K; Strang, M; Alverson, G; Barberis, E; Boeriu, O; Eulisse, G; Govi, G; McCauley, T; Musienko, Y; Muzaffar, S; Osborne, I; Paul, T; Reucroft, S; Swain, J; Taylor, L; Tuura, L; Anastassov, A; Gobbi, B; Kubik, A; Ofierzynski, R A; Pozdnyakov, A; Schmitt, M; Stoynev, S; Velasco, M; Won, S; Antonelli, L; Berry, D; Hildreth, M; Jessop, C; Karmgard, D J; Kolberg, T; Lannon, K; Lynch, S; Marinelli, N; Morse, D M; Ruchti, R; Slaunwhite, J; Warchol, J; Wayne, M; Bylsma, B; Durkin, L S; Gilmore, J; Gu, J; Killewald, P; Ling, T Y; Williams, G; Adam, N; Berry, E; Elmer, P; Garmash, A; Gerbaudo, D; Halyo, V; Hunt, A; Jones, J; Laird, E; Marlow, D; Medvedeva, T; Mooney, M; Olsen, J; Piroué, P; Stickland, D; Tully, C; Werner, J S; Wildish, T; Xie, Z; Zuranski, A; Acosta, J G; Bonnett Del Alamo, M; Huang, X T; Lopez, A; Mendez, H; Oliveros, S; Ramirez Vargas, J E; Santacruz, N; Zatzerklyany, A; Alagoz, E; Antillon, E; Barnes, V E; Bolla, G; Bortoletto, D; Everett, A; Garfinkel, A F; Gecse, Z; Gutay, L; Ippolito, N; Jones, M; Koybasi, O; Laasanen, A T; Leonardo, N; Liu, C; Maroussov, V; Merkel, P; Miller, D H; Neumeister, N; Sedov, A; Shipsey, I; Yoo, H D; Zheng, Y; Jindal, P; Parashar, N; Cuplov, V; Ecklund, K M; Geurts, F J M; Liu, J H; Maronde, D; Matveev, M; Padley, B P; Redjimi, R; Roberts, J; Sabbatini, L; Tumanov, A; Betchart, B; Bodek, A; Budd, H; Chung, Y S; de Barbaro, P; Demina, R; Flacher, H; Gotra, Y; Harel, A; Korjenevski, S; Miner, D C; Orbaker, D; Petrillo, G; Vishnevskiy, D; Zielinski, M; Bhatti, A; Demortier, L; Goulianos, K; Hatakeyama, K; Lungu, G; Mesropian, C; Yan, M; Atramentov, O; Bartz, E; Gershtein, Y; Halkiadakis, E; Hits, D; Lath, A; Rose, K; Schnetzer, S; Somalwar, S; Stone, R; Thomas, S; Watts, T L; Cerizza, G; Hollingsworth, M; Spanier, S; Yang, Z C; York, A; Asaadi, J; Aurisano, A; Eusebi, R; Golyash, A; Gurrola, A; Kamon, T; Nguyen, C N; Pivarski, J; Safonov, A; Sengupta, S; Toback, D; Weinberger, M; Akchurin, N; Berntzon, L; Gumus, K; Jeong, C; Kim, H; Lee, S W; Popescu, S; Roh, Y; Sill, A; Volobouev, I; Washington, E; Wigmans, R; Yazgan, E; Engh, D; Florez, C; Johns, W; Pathak, S; Sheldon, P; Andelin, D; Arenton, M W; Balazs, M; Boutle, S; Buehler, M; Conetti, S; Cox, B; Hirosky, R; Ledovskoy, A; Neu, C; Phillips II, D; Ronquest, M; Yohay, R; Gollapinni, S; Gunthoti, K; Harr, R; Karchin, P E; Mattson, M; Sakharov, A; Anderson, M; Bachtis, M; Bellinger, J N; Carlsmith, D; Crotty, I; Dasu, S; Dutta, S; Efron, J; Feyzi, F; Flood, K; Gray, L; Grogg, K S; Grothe, M; Hall-Wilton, R; Jaworski, M; Klabbers, P; Klukas, J; Lanaro, A; Lazaridis, C; Leonard, J; Loveless, R; Magrans de Abril, M; Mohapatra, A; Ott, G; Polese, G; Reeder, D; Savin, A; Smith, W H; Sourkov, A; Swanson, J; Weinberg, M; Wenman, D; Wensveen, M; White, A

    2010-01-01

    This paper describes the calibration procedure for the drift tubes of the CMS barrel muon system and reports the main results obtained with data collected during a high statistics cosmic ray data-taking period. The main goal of the calibration is to determine, for each drift cell, the minimum time delay for signals relative to the trigger, accounting for the drift velocity within the cell. The accuracy of the calibration procedure is influenced by the random arrival time of cosmic muons. A more refined analysis of the drift velocity was performed during the offline reconstruction phase, which takes into account this feature of cosmic ray events.

  4. A global model for residential energy use: Uncertainty in calibration to regional data

    International Nuclear Information System (INIS)

    van Ruijven, Bas; van Vuuren, Detlef P.; de Vries, Bert; van der Sluijs, Jeroen P.

    2010-01-01

    Uncertainties in energy demand modelling allow for the development of different models, but also leave room for different calibrations of a single model. We apply an automated model calibration procedure to analyse calibration uncertainty of residential sector energy use modelling in the TIMER 2.0 global energy model. This model simulates energy use on the basis of changes in useful energy intensity, technology development (AEEI) and price responses (PIEEI). We find that different implementations of these factors yield behavioural model results. Model calibration uncertainty is identified as influential source for variation in future projections: amounting 30% to 100% around the best estimate. Energy modellers should systematically account for this and communicate calibration uncertainty ranges. (author)

  5. Calibration of communication skills items in OSCE checklists according to the MAAS-Global.

    Science.gov (United States)

    Setyonugroho, Winny; Kropmans, Thomas; Kennedy, Kieran M; Stewart, Brian; van Dalen, Jan

    2016-01-01

    Communication skills (CS) are commonly assessed using 'communication items' in Objective Structured Clinical Examination (OSCE) station checklists. Our aim is to calibrate the communication component of OSCE station checklists according to the MAAS-Global which is a valid and reliable standard to assess CS in undergraduate medical education. Three raters independently compared 280 checklists from 4 disciplines contributing to the undergraduate year 4 OSCE against the 17 items of the MAAS-Global standard. G-theory was used to analyze the reliability of this calibration procedure. G-Kappa was 0.8. For two raters G-Kappa is 0.72 and it fell to 0.57 for one rater. 46% of the checklist items corresponded to section three of the MAAS-Global (i.e. medical content of the consultation), whilst 12% corresponded to section two (i.e. general CS), and 8.2% to section one (i.e. CS for each separate phase of the consultation). 34% of the items were not considered to be CS. A G-Kappa of 0.8 confirms a reliable and valid procedure for calibrating OSCE CS checklist items using the MAAS-Global. We strongly suggest that such a procedure is more widely employed to arrive at a stable (valid and reliable) judgment of the communication component in existing checklists for medical students' communication behaviours. It is possible to measure the 'true' caliber of CS in OSCE stations. Students' results are thereby comparable between and across stations, students and institutions. A reliable calibration procedure requires only two raters. Copyright © 2015. Published by Elsevier Ireland Ltd.

  6. The Self-Calibration Test of flowmeter installed in STELLA(Sodium Integral Effect Test Loop for Safety Simulation and Assessment) facility

    Energy Technology Data Exchange (ETDEWEB)

    Jung, Minhwan; Jeong, Ji-Young [Korea Atomic Energy Research Institute, Daejeon (Korea, Republic of)

    2015-10-15

    The objective of this study is to describe the procedure of the self-calibration test for the flowmeters and to analyze the result of the test. In this work, the test procedure of the self-calibration of two flowmeters (FT-101, FT-102) installed in STELLA facility was described and the test result was analyzed. In regard to the long-term SFR development plan, a large-scale sodium thermal-hydraulic test project is being progressed by KAERI. This project is called STELLA (Sodium Integral Effect Test Loop for Safety Simulation and Assessment), and it is proceeding by adopting the QA (Quality Assurance) program. Due to the specificity of an experiment using sodium(Na) categorized as Class 3(pyrophoric material and water-prohibiting substance) by the Safety Control of Dangerous Substances Act, it is necessary to apply QA in consideration of the sodium experiment environment in certain parts. The one of them is about calibration of measuring instrument such as a flowmeter, thermocouple and pressure gauge. It is described in the QAP (Quality Assurance Procedures) of KAERI that calibration work should be conducted in accordance with self-calibration procedures in a special case where conventional calibration is not practicable. The calibration of two flowmeters (FT-101, FT-102) installed in STELLA facility is the typical example. As a result of test, it was confirmed that the flowmeters meet the pass criterion. Therefore, it was concluded that the flowmeters maintain instrument capacity a year ago.

  7. Transfer standard device to improve the traceable calibration of physiotherapy ultrasound machines

    NARCIS (Netherlands)

    Hekkenberg, R.T.; Richards, A.; Beissner, K.; Zeqiri, B.; Bezemer, R.A.; Hodnett, M.; Prout, G.; Cantrall, C.

    2006-01-01

    Ultrasound (US) physiotherapy as a clinical treatment is extremely common in the Western world. Internationally, regulation to ensure safe application of US physiotherapy by regular calibration ranges from nil to mandatory. The need for a portable power standard (PPS) has been addressed within a

  8. Certification of flow measurement methods and calibration of flowmeters in oil and gas industry using radioactive tracers

    Energy Technology Data Exchange (ETDEWEB)

    Brandao, L. E. E.; Kenup, H.O. [Instituto de Engenharia Nuclear (IEN/CNEN-RJ), Rio de Janeiro, RJ (Brazil); Nóbrega, Armi W.; Gonçalves, E.R.; Dualibi Filho, J.C.; Malheiro, R.L., E-mail: brandaos@ien.gov.br, E-mail: atomum@atomum.com.br [ATOMUM Serviços Tecnológicos Ltda, Rio de Janeiro, RJ (Brazil)

    2017-07-01

    Oil and gas, mineral and water supply plants are the typical example of industrial facilities where they need accuracy in fluid flow measurement's procedures. In these installations, large quantities of materials are moving daily inside pipes, and one of the major problems of these industries is the safety of operations in transport of these materials. In order to monitor the transference processes many flowmeters are installed in the pipes to measure constantly the condition of the transported fluid. These flowmeters must be periodically calibrated, and one of the problems in this standardization is that these instruments need to be removed from the pipelines where they are operating and transported to the accredited laboratories to be calibrated. To remove a flowmeter from a pipe is not an easy operation, and in most of the cases, it is a very expensive operation, and in addition, any changes in the pipeline configuration can provoke leaks and when a leak occurs, it is impossible to operate the whole line and provokes accidents. Radiotracer technique is noninvasive fluid flow measure and using proper radiotracers is possible to measure organic, aqueous and gaseous flows without any disturbance in the normal operation of the pipeline. In this work, we present the ATOMUM{sub T}RACER, a software used to determine geometrical and technical procedures to calculate the Residence Time Curves using radiotracer techniques to accurately measure flow of fluids in pipelines with uncertainties around 0.7%. (author)

  9. Certification of flow measurement methods and calibration of flowmeters in oil and gas industry using radioactive tracers

    International Nuclear Information System (INIS)

    Brandao, L. E. E.; Kenup, H.O.; Nóbrega, Armi W.; Gonçalves, E.R.; Dualibi Filho, J.C.; Malheiro, R.L.

    2017-01-01

    Oil and gas, mineral and water supply plants are the typical example of industrial facilities where they need accuracy in fluid flow measurement's procedures. In these installations, large quantities of materials are moving daily inside pipes, and one of the major problems of these industries is the safety of operations in transport of these materials. In order to monitor the transference processes many flowmeters are installed in the pipes to measure constantly the condition of the transported fluid. These flowmeters must be periodically calibrated, and one of the problems in this standardization is that these instruments need to be removed from the pipelines where they are operating and transported to the accredited laboratories to be calibrated. To remove a flowmeter from a pipe is not an easy operation, and in most of the cases, it is a very expensive operation, and in addition, any changes in the pipeline configuration can provoke leaks and when a leak occurs, it is impossible to operate the whole line and provokes accidents. Radiotracer technique is noninvasive fluid flow measure and using proper radiotracers is possible to measure organic, aqueous and gaseous flows without any disturbance in the normal operation of the pipeline. In this work, we present the ATOMUM T RACER, a software used to determine geometrical and technical procedures to calculate the Residence Time Curves using radiotracer techniques to accurately measure flow of fluids in pipelines with uncertainties around 0.7%. (author)

  10. When to Make Mountains out of Molehills: The Pros and Cons of Simple and Complex Model Calibration Procedures

    Science.gov (United States)

    Smith, K. A.; Barker, L. J.; Harrigan, S.; Prudhomme, C.; Hannaford, J.; Tanguy, M.; Parry, S.

    2017-12-01

    Earth and environmental models are relied upon to investigate system responses that cannot otherwise be examined. In simulating physical processes, models have adjustable parameters which may, or may not, have a physical meaning. Determining the values to assign to these model parameters is an enduring challenge for earth and environmental modellers. Selecting different error metrics by which the models results are compared to observations will lead to different sets of calibrated model parameters, and thus different model results. Furthermore, models may exhibit `equifinal' behaviour, where multiple combinations of model parameters lead to equally acceptable model performance against observations. These decisions in model calibration introduce uncertainty that must be considered when model results are used to inform environmental decision-making. This presentation focusses on the uncertainties that derive from the calibration of a four parameter lumped catchment hydrological model (GR4J). The GR models contain an inbuilt automatic calibration algorithm that can satisfactorily calibrate against four error metrics in only a few seconds. However, a single, deterministic model result does not provide information on parameter uncertainty. Furthermore, a modeller interested in extreme events, such as droughts, may wish to calibrate against more low flows specific error metrics. In a comprehensive assessment, the GR4J model has been run with 500,000 Latin Hypercube Sampled parameter sets across 303 catchments in the United Kingdom. These parameter sets have been assessed against six error metrics, including two drought specific metrics. This presentation compares the two approaches, and demonstrates that the inbuilt automatic calibration can outperform the Latin Hypercube experiment approach in single metric assessed performance. However, it is also shown that there are many merits of the more comprehensive assessment, which allows for probabilistic model results, multi

  11. Review of technological advancements in calibration systems for laser vision correction

    Science.gov (United States)

    Arba-Mosquera, Samuel; Vinciguerra, Paolo; Verma, Shwetabh

    2018-02-01

    Using PubMed and our internal database, we extensively reviewed the literature on the technological advancements in calibration systems, with a motive to present an account of the development history, and latest developments in calibration systems used in refractive surgery laser systems. As a second motive, we explored the clinical impact of the error introduced due to the roughness in ablation and its corresponding effect on system calibration. The inclusion criterion for this review was strict relevance to the clinical questions under research. The existing calibration methods, including various plastic models, are highly affected by various factors involved in refractive surgery, such as temperature, airflow, and hydration. Surface roughness plays an important role in accurate measurement of ablation performance on calibration materials. The ratio of ablation efficiency between the human cornea and calibration material is very critical and highly dependent on the laser beam characteristics and test conditions. Objective evaluation of the calibration data and corresponding adjustment of the laser systems at regular intervals are essential for the continuing success and further improvements in outcomes of laser vision correction procedures.

  12. Calibration and monitoring of the ATLAS Tile calorimeter

    CERN Document Server

    Boumediene, Djamel Eddine; The ATLAS collaboration

    2017-01-01

    The ATLAS Tile Calorimeter (TileCal) is the central section of the hadronic calorimeter of the ATLAS experiment and provides important information for reconstruction of hadrons, jets, hadronic decays of tau leptons and missing transverse energy. This sampling calorimeter uses steel plates as absorber and scintillating tiles as active medium. The light produced by the passage of charged particles is transmitted by wavelength shifting fibres to photomultiplier tubes (PMTs). PMT signals are then digitized at 40~MHz and stored on detector and are only transferred off detector once the first level trigger acceptance has been confirmed. The readout is segmented into about 5000 cells (longitudinally and transversally), each of them being read out by two PMTs in parallel. To calibrate and monitor the stability and performance of each part of the readout chain, a set of calibration systems is used. The TileCal calibration system comprises Cesium radioactive sources, laser, charge injection elements and an integrator b...

  13. Calibration of dosimeters used in radiotherapy. A manual

    International Nuclear Information System (INIS)

    1994-01-01

    The present manual is a revision of IAEA Technical Reports Series No.185, published in 1979. This manual is intended for use by the network of Secondary Standard Dosimetry Laboratories (SSDLs) which was set up by the IAEA and the World Health Organization (WHO). The objectives of the SSDL network are to: calibrate radiation dosimeters and ancillary instruments; serve as a link between primary standard laboratories and radiation users; and provide advice and assistance in all aspects of radiation metrology. The various calibration procedures are described, their relative advantages and disadvantages are discussed, and criteria are put forward to help an SSDL decide which procedure is the best to use in order to meet a particular requirement. The information in this manual should also be of value to other similar laboratories, usually associated with hospitals, which are not formally part of the IAEA/WHO network of SSDLs. 26 refs, 6 figs, 6 tabs

  14. 40 CFR 92.116 - Engine output measurement system calibrations.

    Science.gov (United States)

    2010-07-01

    ... manufacturer's data, actual measurement, or the value recorded from the previous calibration used for this... load-cell or transfer standard may be used to verify the in-use torque measurement system. (i) The... spaced torque values as indicated by the master load-cell for each in-use range used. (v) The in-use...

  15. Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear

    Energy Technology Data Exchange (ETDEWEB)

    Cortes P, A; Garcia D, O C; Ortiz P, I; Becerril V, A [Instituto nacional de Investigaciones Nucleares, Laboratorio de Patrones Radiactivos, Departamento de Metrologia, A.P. 18-1027. C.P. 11801 Mexico D.F. (Mexico)

    2000-07-01

    With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some {mu} Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some {mu} Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm{sup 3} contained in polyethylene small bottles of 25 cm{sup 3} elaborated in the LPR and whose activity is measured with respect to the National Patron ININ-PNM-2. The sources of mid-large life are ampoules with 5 cm{sup 3} of radioactive solution, calibrated in activity by the primary laboratory, LNHB (formerly LMRI) of France. (Author)

  16. Calibration of three-axis magnetometers with differential evolution algorithm

    International Nuclear Information System (INIS)

    Pang, Hongfeng; Zhang, Qi; Wang, Wei; Wang, Junya; Li, Ji; Luo, Shitu; Wan, Chengbiao; Chen, Dixiang; Pan, Mengchun; Luo, Feilu

    2013-01-01

    The accuracy of three-axis magnetometers is influenced by different scale and bias of each axis and nonorthogonality between axes. One limitation of traditional iteration methods is that initial parameters influence the calibration, thus leading to the local optimal or wrong results. In this paper, a new method is proposed to calibrate three-axis magnetometers. To employ this method, a nonmagnetic rotation platform, a proton magnetometer, a DM-050 three-axis magnetometer and the differential evolution (DE) algorithm are used. The performance of this calibration method is analyzed with simulation and experiment. In simulation, the calibration results of DE, unscented Kalman filter (UKF), recursive least squares (RLS) and genetic algorithm (GA) are compared. RMS error using DE is least, which is reduced from 81.233 nT to 1.567 nT. Experimental results show that comparing with UKF, RLS and GA, the DE algorithm has not only the least calibration error but also the best robustness. After calibration, RMS error is reduced from 68.914 nT to 2.919 nT. In addition, the DE algorithm is not sensitive to initial parameters, which is an important advantage compared with traditional iteration algorithms. The proposed algorithm can avoid the troublesome procedure to select suitable initial parameters, thus it can improve the calibration performance of three-axis magnetometers. - Highlights: • The calibration results and robustness of UKF, GA, RLS and DE algorithm are analyzed. • Calibration error of DE is the least in simulation and experiment. • Comparing with traditional calibration algorithms, DE is not sensitive to initial parameters. • It can improve the calibration performance of three-axis magnetometers

  17. The analytical calibration in (bio)imaging/mapping of the metallic elements in biological samples--definitions, nomenclature and strategies: state of the art.

    Science.gov (United States)

    Jurowski, Kamil; Buszewski, Bogusław; Piekoszewski, Wojciech

    2015-01-01

    Nowadays, studies related to the distribution of metallic elements in biological samples are one of the most important issues. There are many articles dedicated to specific analytical atomic spectrometry techniques used for mapping/(bio)imaging the metallic elements in various kinds of biological samples. However, in such literature, there is a lack of articles dedicated to reviewing calibration strategies, and their problems, nomenclature, definitions, ways and methods used to obtain quantitative distribution maps. The aim of this article was to characterize the analytical calibration in the (bio)imaging/mapping of the metallic elements in biological samples including (1) nomenclature; (2) definitions, and (3) selected and sophisticated, examples of calibration strategies with analytical calibration procedures applied in the different analytical methods currently used to study an element's distribution in biological samples/materials such as LA ICP-MS, SIMS, EDS, XRF and others. The main emphasis was placed on the procedures and methodology of the analytical calibration strategy. Additionally, the aim of this work is to systematize the nomenclature for the calibration terms: analytical calibration, analytical calibration method, analytical calibration procedure and analytical calibration strategy. The authors also want to popularize the division of calibration methods that are different than those hitherto used. This article is the first work in literature that refers to and emphasizes many different and complex aspects of analytical calibration problems in studies related to (bio)imaging/mapping metallic elements in different kinds of biological samples. Copyright © 2014 Elsevier B.V. All rights reserved.

  18. Calibration of proportional counters in microdosimetry

    International Nuclear Information System (INIS)

    Varma, M.N.

    1982-01-01

    Many microdosimetric spectra for low LET as well as high LET radiations are measured using commercially available (similar to EG and G) Rossi proportional counters. This paper discusses the corrections to be applied to data when calibration of the counter is made using one type of radiation, and then the counter is used in a different radiation field. The principal correction factor is due to differences in W-value of the radiation used for calibration and the radiation for which microdosimetric measurements are made. Both propane and methane base tissue-equivalent (TE) gases are used in these counters. When calibrating the detectors, it is important to use the correct stopping power value for that gas. Deviations in y-bar/sub F/ and y-bar/sub D/ are calculated for 60 Co using different extrapolation procedures from 0.15 keV/μm to zero event size. These deviations can be as large as 30%. Advantages of reporting microdosimetric parameters such as y-bar/sub F/ and y-bar/sub D/ above a certain minimum cut-off are discussed

  19. The study of new calibration features in the Harshaw TLD system

    International Nuclear Information System (INIS)

    Luo, L. Z.

    2007-01-01

    The Harshaw TLD system has three key calibration procedures: the Reader, the Dosemeter and the Algorithm. These functions must be properly calibrated for the system to achieve the optimum results. For the conventional reader and dosemeter calibration, Harshaw TLD recommends a pre-fade and a post-fade of 24-48 h when calibrating the system for LiF:Mg,Ti type dosemeter. It is also recommended that keeping the fade time consistent is important to maintain the quality of the system performance. In recent years, new calibration features have been introduced into the Harshaw TLD models 6600 and 8800 operating systems. These new features are Auto Calibration, Auto QC and Auto Blank, and they give the user the ability to set up the clear-expose-read process to be performed automatically in a sequence for each dosemeter. This saves processing time and keeps the fade time the same. However, since the fade time is near zero, will it affect the TLD system calibration factors? What should the user expect? This paper presents a study of the effect of Auto Calibration/Auto QC to the TLD operation. (authors)

  20. Application of heuristic and machine-learning approach to engine model calibration

    Science.gov (United States)

    Cheng, Jie; Ryu, Kwang R.; Newman, C. E.; Davis, George C.

    1993-03-01

    Automation of engine model calibration procedures is a very challenging task because (1) the calibration process searches for a goal state in a huge, continuous state space, (2) calibration is often a lengthy and frustrating task because of complicated mutual interference among the target parameters, and (3) the calibration problem is heuristic by nature, and often heuristic knowledge for constraining a search cannot be easily acquired from domain experts. A combined heuristic and machine learning approach has, therefore, been adopted to improve the efficiency of model calibration. We developed an intelligent calibration program called ICALIB. It has been used on a daily basis for engine model applications, and has reduced the time required for model calibrations from many hours to a few minutes on average. In this paper, we describe the heuristic control strategies employed in ICALIB such as a hill-climbing search based on a state distance estimation function, incremental problem solution refinement by using a dynamic tolerance window, and calibration target parameter ordering for guiding the search. In addition, we present the application of a machine learning program called GID3* for automatic acquisition of heuristic rules for ordering target parameters.

  1. Quality assurance programme at the National Calibration Laboratory in Tanzania

    International Nuclear Information System (INIS)

    Muhogora, W.E.; Yoloye, O.; Ngaile, J.; Lema, U.S.

    2000-01-01

    A quality assurance programme at the National Calibration Laboratory for ionizing radiation in Tanzania is described. The programme focuses mainly on regular stability check source and reference output measurements, performance testing of TLD systems as well as some external audit checks. It is found that the stability check source measurements are within ± 1%. Similarly, the air kerma rate measurements agree well with calibration uncertainties, that is ± 2% for protection level measurements and ± 1.5% for clinical dosimetry. The results of comparison of dose measurements done on site and those obtained from some external audit checks are also within requirements. This shows that the working standards have been kept with good care, and that the traceability to the international measurement system is adequately maintained. Some examples on calibration transfer activities are briefly discussed

  2. Tritium monitor calibration at Los Alamos National Laboratory

    International Nuclear Information System (INIS)

    Bjork, C.J.; Aikin, D.J.; Houlton, T.W.

    1997-08-01

    Tritium in air is monitored at Los Alamos National Laboratory (LANL) with air breathing instruments based on ionization chambers. Stack emissions are continuously monitored from sample tubes which each connect to a Tritium bubble which differentially collects HTO and HT. A set of glass vials of glycol capture the HTO. The HT is oxidized with a palladium catalyst and the resultant HTO is captured in a second set of vials of glycol. The glycol is counted with a liquid scintillation counter. All calibrations are performed with tritium containing gas. The Radiation Instrumentation and Calibration (RIC) Team has constructed and maintains two closed loop gas handling systems based on femto TECH model U24 tritium ion chamber monitors: a fixed system housed in a fume hood and a portable system mounted on two two wheeled hand trucks. The U24 monitors are calibrated against tritium in nitrogen gas standards. They are used as standard transfer instruments to calibrate other ion chamber monitors with tritium in nitrogen, diluted with air. The gas handling systems include a circulation pump which permits a closed circulation loop to be established among the U24 monitor and typically two to four other monitors of a given model during calibration. Fixed and portable monitors can be calibrated. The stack bubblers are calibrated in the field by: blending a known concentration of tritium in air within the known volume of the two portable carts, coupled into a common loop; releasing that gas mixture into a ventilation intake to the stack; collecting oxidized tritium in the bubbler; counting the glycol; and using the stack and bubbler flow rates, computing the bubbler's efficiency. Gas calibration has become a convenient and quality tool in maintaining the tritium monitors at LANL

  3. Phase calibration strategies for synchrotron RF signals

    Energy Technology Data Exchange (ETDEWEB)

    Andreev, Aleksandr [TEMF, Technische Universitaet Darmstadt (Germany); Klingbeil, Harald [TEMF, Technische Universitaet Darmstadt (Germany); GSI Helmholtzzentrum fuer Schwerionenforschung GmbH, Darmstadt (Germany); Lens, Dieter [GSI Helmholtzzentrum fuer Schwerionenforschung GmbH, Darmstadt (Germany)

    2016-07-01

    For the FAIR facility that is currently under construction, the beam quality requirements impose several demands on the low-level RF (LLRF) systems. For example the phase error of the gap voltage of a specific RF cavity must be less than 1 . The RF reference signals for the FAIR synchrotron RF cavity systems are generated by direct digital synthesis modules (DDS) mounted in one crate called Group-DDS. In order to allow performing various multi-harmonic operations, each DDS unit operates at a certain mode defined by the harmonic number that can be changed during the operation. Since the DDS modules generate reference RF signals for different LLRF systems, the precise calibration of units to compensate the different phase response is of importance. The currently used calibration procedure is done with a fixed harmonic number for each module and uses the DDS module configured to the highest harmonic number as a reference. If the harmonic number of the DDS module is changed, one then has to repeat the calibration for the new values. Therefore, a new calibration method with respect to the absolute phases of DDS modules is under development and will be presented.

  4. Lateral force calibration in atomic force microscopy: A new lateral force calibration method and general guidelines for optimization

    International Nuclear Information System (INIS)

    Cannara, Rachel J.; Eglin, Michael; Carpick, Robert W.

    2006-01-01

    Proper force calibration is a critical step in atomic and lateral force microscopies (AFM/LFM). The recently published torsional Sader method [C. P. Green et al., Rev. Sci. Instrum. 75, 1988 (2004)] facilitates the calculation of torsional spring constants of rectangular AFM cantilevers by eliminating the need to obtain information or make assumptions regarding the cantilever's material properties and thickness, both of which are difficult to measure. Complete force calibration of the lateral signal in LFM requires measurement of the lateral signal deflection sensitivity as well. In this article, we introduce a complete lateral force calibration procedure that employs the torsional Sader method and does not require making contact between the tip and any sample. In this method, a colloidal sphere is attached to a 'test' cantilever of the same width, but different length and material as the 'target' cantilever of interest. The lateral signal sensitivity is calibrated by loading the colloidal sphere laterally against a vertical sidewall. The signal sensitivity for the target cantilever is then corrected for the tip length, total signal strength, and in-plane bending of the cantilevers. We discuss the advantages and disadvantages of this approach in comparison with the other established lateral force calibration techniques, and make a direct comparison with the 'wedge' calibration method. The methods agree to within 5%. The propagation of errors is explicitly considered for both methods and the sources of disagreement discussed. Finally, we show that the lateral signal sensitivity is substantially reduced when the laser spot is not centered on the detector

  5. Challenges in X-band Weather Radar Data Calibration

    DEFF Research Database (Denmark)

    Thorndahl, Søren; Rasmussen, Michael R.

    2009-01-01

    Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis and real time control purposes. In these contexts, it is all-important that the radar data well calibrated and adjusted in order to obtain valid quantitative precipitation e...... estimates. This paper compares two calibration procedures for a small marine X-band radar by comparing radar data with rain gauge data. Validation shows a very good consensus with regards to precipitation volumes, but more diverse results on peak rain intensities.......Application of weather radar data in urban hydrology is evolving and radar data is now applied for both modelling, analysis and real time control purposes. In these contexts, it is all-important that the radar data well calibrated and adjusted in order to obtain valid quantitative precipitation...

  6. Study on routines and procedures of calibration of clinical dosemeter on gamma radiation beams at the IPEN-CNEN/SP, Brazil

    International Nuclear Information System (INIS)

    Damatto, Willian B.; Potiens, Maria da Penha A.; Santos, Gelson P.; Vivolo, Vitor

    2009-01-01

    The calibration consists in the comparison of the equipment to be calibrated with other equipment already calibrated by another standard, which means the equipment that will accomplish shall be considered secondary or tertiary, depending on the calibration that will be accomplished. As the clinical dosemeters are used for quality control services in the hospitals, these dosemeters need a reliable measurement precision. Therefore, this work intends to demonstrate the importance of the clinical dosemeter and the calibration, requiring special care face to all clinical set

  7. Calibrated expressions for welding and their application to isotherm width in a thick plate

    Directory of Open Access Journals (Sweden)

    Gentry Wood

    2014-09-01

    Full Text Available The present paper introduces a possible solution to the limitations of modern trial and error solutions to welding procedure development. The difficulties of finding generalized solutions to Rosenthal´s equation are discussed and the Minimal Representation and Calibration approach is introduced as a promising procedure for developing these solutions. Dominant factors are identified, with effects from secondary phenomena being taken into account by correction factors. These correction factors are then calibrated and presented in a form that can be easily computed, and therefore be amendable to industry. The approach is then demonstrated by determining the isotherm width from Rosenthal´s thick plate solution. Comparison of the calibrated scaling equations to Rosenthal´s exact solution showed a maximum error of less than 0.8% for any isotherm.

  8. Hierarchical calibration and validation framework of bench-scale computational fluid dynamics simulations for solvent-based carbon capture. Part 2: Chemical absorption across a wetted wall column: Original Research Article: Hierarchical calibration and validation framework of bench-scale computational fluid dynamics simulations

    Energy Technology Data Exchange (ETDEWEB)

    Wang, Chao [Pacific Northwest National Laboratory, Physical and Computational Sciences Directorate, Richland WA; Xu, Zhijie [Pacific Northwest National Laboratory, Physical and Computational Sciences Directorate, Richland WA; Lai, Kevin [Pacific Northwest National Laboratory, Physical and Computational Sciences Directorate, Richland WA; Whyatt, Greg [Energy and Environment Directorate, Pacific Northwest National Laboratory, Richland WA; Marcy, Peter W. [Los Alamos National Laboratory, Statistical Sciences Group, Los Alamos NM; Sun, Xin [Oak Ridge National Laboratory, Energy and Transportation Science Division, Oak Ridge TN

    2017-10-24

    The first part of this paper (Part 1) presents a numerical model for non-reactive physical mass transfer across a wetted wall column (WWC). In Part 2, we improved the existing computational fluid dynamics (CFD) model to simulate chemical absorption occurring in a WWC as a bench-scale study of solvent-based carbon dioxide (CO2) capture. To generate data for WWC model validation, CO2 mass transfer across a monoethanolamine (MEA) solvent was first measured on a WWC experimental apparatus. The numerical model developed in this work has the ability to account for both chemical absorption and desorption of CO2 in MEA. In addition, the overall mass transfer coefficient predicted using traditional/empirical correlations is conducted and compared with CFD prediction results for both steady and wavy falling films. A Bayesian statistical calibration algorithm is adopted to calibrate the reaction rate constants in chemical absorption/desorption of CO2 across a falling film of MEA. The posterior distributions of the two transport properties, i.e., Henry’s constant and gas diffusivity in the non-reacting nitrous oxide (N2O)/MEA system obtained from Part 1 of this study, serves as priors for the calibration of CO2 reaction rate constants after using the N2O/CO2 analogy method. The calibrated model can be used to predict the CO2 mass transfer in a WWC for a wider range of operating conditions.

  9. Spectral Transfer Learning using Information Geometry for a User-Independent Brain-Computer Interface

    Directory of Open Access Journals (Sweden)

    Nicholas Roy Waytowich

    2016-09-01

    Full Text Available Recent advances in signal processing and machine learning techniques have enabled the application of Brain-Computer Interface (BCI technologies to fields such as medicine, industry and recreation. However, BCIs still suffer from the requirement of frequent calibration sessions due to the intra- and inter- individual variability of brain-signals, which makes calibration suppression through transfer learning an area of increasing interest for the development of practical BCI systems. In this paper, we present an unsupervised transfer method (spectral transfer using information geometry, STIG, which ranks and combines unlabeled predictions from an ensemble of information geometry classifiers built on data from individual training subjects. The STIG method is validated in both offline and real-time feedback analysis during a rapid serial visual presentation task (RSVP. For detection of single-trial, event-related potentials (ERPs, the proposed method can significantly outperform existing calibration-free techniques as well as outperform traditional within-subject calibration techniques when limited data is available. This method demonstrates that unsupervised transfer learning for single-trial detection in ERP-based BCIs can be achieved without the requirement of costly training data, representing a step-forward in the overall goal of achieving a practical user-independent BCI system.

  10. Automatic Chessboard Detection for Intrinsic and Extrinsic Camera Parameter Calibration

    Directory of Open Access Journals (Sweden)

    Jose María Armingol

    2010-03-01

    Full Text Available There are increasing applications that require precise calibration of cameras to perform accurate measurements on objects located within images, and an automatic algorithm would reduce this time consuming calibration procedure. The method proposed in this article uses a pattern similar to that of a chess board, which is found automatically in each image, when no information regarding the number of rows or columns is supplied to aid its detection. This is carried out by means of a combined analysis of two Hough transforms, image corners and invariant properties of the perspective transformation. Comparative analysis with more commonly used algorithms demonstrate the viability of the algorithm proposed, as a valuable tool for camera calibration.

  11. Design and calibration of a scanning tunneling microscope for large machined surfaces

    Energy Technology Data Exchange (ETDEWEB)

    Grigg, D.A.; Russell, P.E.; Dow, T.A.

    1988-12-01

    During the last year the large sample STM has been designed, built and used for the observation of several different samples. Calibration of the scanner for prope dimensional interpretation of surface features has been a chief concern, as well as corrections for non-linear effects such as hysteresis during scans. Several procedures used in calibration and correction of piezoelectric scanners used in the laboratorys STMs are described.

  12. High-accuracy self-calibration method for dual-axis rotation-modulating RLG-INS

    Science.gov (United States)

    Wei, Guo; Gao, Chunfeng; Wang, Qi; Wang, Qun; Long, Xingwu

    2017-05-01

    Inertial navigation system has been the core component of both military and civil navigation systems. Dual-axis rotation modulation can completely eliminate the inertial elements constant errors of the three axes to improve the system accuracy. But the error caused by the misalignment angles and the scale factor error cannot be eliminated through dual-axis rotation modulation. And discrete calibration method cannot fulfill requirements of high-accurate calibration of the mechanically dithered ring laser gyroscope navigation system with shock absorbers. This paper has analyzed the effect of calibration error during one modulated period and presented a new systematic self-calibration method for dual-axis rotation-modulating RLG-INS. Procedure for self-calibration of dual-axis rotation-modulating RLG-INS has been designed. The results of self-calibration simulation experiment proved that: this scheme can estimate all the errors in the calibration error model, the calibration precision of the inertial sensors scale factor error is less than 1ppm and the misalignment is less than 5″. These results have validated the systematic self-calibration method and proved its importance for accuracy improvement of dual -axis rotation inertial navigation system with mechanically dithered ring laser gyroscope.

  13. Effect of contact stiffness on wedge calibration of lateral force in atomic force microscopy

    International Nuclear Information System (INIS)

    Wang Fei; Zhao Xuezeng

    2007-01-01

    Quantitative friction measurement of nanomaterials in atomic force microscope requires accurate calibration method for lateral force. The effect of contact stiffness on lateral force calibration of atomic force microscope is discussed in detail and an improved calibration method is presented. The calibration factor derived from the original method increased with the applied normal load, which indicates that separate calibration should be required for every given applied normal load to keep the accuracy of friction measurement. We improve the original method by introducing the contact factor, which is derived from the contact stiffness between the tip and the sample, to the calculation of calibration factors. The improved method makes the calculation of calibration factors under different applied normal loads possible without repeating the calibration procedure. Comparative experiments on a silicon wafer have been done by both the two methods to validate the method in this article

  14. Binary Pseudo-Random Gratings and Arrays for Calibration of Modulation Transfer Functions of Surface Profilometers

    Energy Technology Data Exchange (ETDEWEB)

    Barber, Samuel K.; Anderson, Erik D.; Cambie, Rossana; McKinney, Wayne R.; Takacs, Peter Z.; Stover, John C.; Voronov, Dmitriy L.; Yashchuk, Valeriy V.

    2009-09-11

    A technique for precise measurement of the modulation transfer function (MTF), suitable for characterization of a broad class of surface profilometers, is investigated in detail. The technique suggested in [Proc. SPIE 7077-7, (2007), Opt. Eng. 47(7), 073602-1-5 (2008)]is based on use of binary pseudo-random (BPR) gratings and arrays as standard MTF test surfaces. Unlike most conventional test surfaces, BPR gratings and arrays possess white-noise-like inherent power spectral densities (PSD), allowing the direct determination of the one- and two-dimensional MTF, respectively, with a sensitivity uniform over the entire spatial frequency range of a profiler. In the cited work, a one dimensional realization of the suggested method based on use of BPR gratings has been demonstrated. Here, a high-confidence of the MTF calibration technique is demonstrated via cross comparison measurements of a number of two dimensional BPR arrays using two different interferometric microscopes and a scatterometer. We also present the results of application of the experimentally determined MTF correction to the measurement taken with the MicromapTM-570 interferometric microscope of the surface roughness of a super-polished test mirror. In this particular case, without accounting for the instrumental MTF, the surface rms roughness over half of the instrumental spatial frequency bandwidth would be underestimated by a factor of approximately 1.4.

  15. The Grover energy transfer algorithm for relativistic speeds

    International Nuclear Information System (INIS)

    Garcia-Escartin, Juan Carlos; Chamorro-Posada, Pedro

    2010-01-01

    Grover's algorithm for quantum search can also be applied to classical energy transfer. The procedure takes a system in which the total energy is equally distributed among N subsystems and transfers most of it to one marked subsystem. We show that in a relativistic setting the efficiency of this procedure can be improved. We will consider the transfer of relativistic kinetic energy in a series of elastic collisions. In this case, the number of steps of the energy transfer procedure approaches 1 as the initial velocities of the objects become closer to the speed of light. This is a consequence of introducing nonlinearities in the procedure. However, the maximum attainable transfer will depend on the particular combination of speed and number of objects. In the procedure, we will use N elements, as in the classical non-relativistic case, instead of the log 2 (N) states of the quantum algorithm.

  16. Quartz microbalance device for transfer into ultrahigh vacuum systems

    International Nuclear Information System (INIS)

    Stavale, F.; Achete, C. A.; Niehus, H.

    2008-01-01

    An uncomplicated quartz microbalance device has been developed which is transferable into ultrahigh vacuum (UHV) systems. The device is extremely useful for flux calibration of different kinds of material evaporators. Mounted on a commercial specimen holder, the device allows fast quartz microbalance transfer into the UHV and subsequent positioning exactly to the sample location where subsequent thin film deposition experiments shall be carried out. After backtransfer into an UHV sample stage, the manipulator may be loaded in situ with the specimen suited for the experiment. The microbalance device capability is demonstrated for monolayer and submonolayer vanadium depositions with an achieved calibration sensitivity of less the 0.001 ML coverage.

  17. 21 CFR 58.81 - Standard operating procedures.

    Science.gov (United States)

    2010-04-01

    ...) Data handling, storage, and retrieval. (11) Maintenance and calibration of equipment. (12) Transfer... LABORATORY PRACTICE FOR NONCLINICAL LABORATORY STUDIES Testing Facilities Operation § 58.81 Standard... forth nonclinical laboratory study methods that management is satisfied are adequate to insure the...

  18. Calibration of radioprotection equipment gamma radiation at the Laboratory of Ionizing Radiation Metrology - DEN/UFPE

    International Nuclear Information System (INIS)

    Nazario, Macilene; Khoury, Helen; Hazin, Clovis

    2003-01-01

    This work presents aspects of the radioprotection equipment calibration service of the Laboratory for Metrology of Ionizing Radiations (LMRI) of the DEN/UFPE related to the calibration procedures, characteristics of the radiation beam and the evaluation of equipment calibrated in the period of 2001-2002. The LMRI-DEN/UFPE is one of the four laboratories in Brazil licensed by the Brazilian Nuclear Energy Commission for the execution of calibration services on area, surface contamination and personal monitors used by industries, hospitals, universities and research institutes using radioactive sources

  19. MP.EXE, a Calculation Program for Pressure Reciprocity Calibration of Microphones

    DEFF Research Database (Denmark)

    Rasmussen, Knud

    1998-01-01

    A computer program is described which calculates the pressure sensitivity of microphones based on measurements of the electrical transfer impedance in a reciprocity calibration set-up. The calculations are performed according to the International Standard IEC 6194-2. In addition a number of options...

  20. 40 CFR 86.1217-96 - Evaporative emission enclosure calibrations.

    Science.gov (United States)

    2010-07-01

    ... procedure: (1) Carefully measure the internal length, width and height of the enclosure, accounting for... 40 Protection of Environment 19 2010-07-01 2010-07-01 false Evaporative emission enclosure calibrations. 86.1217-96 Section 86.1217-96 Protection of Environment ENVIRONMENTAL PROTECTION AGENCY...

  1. Smart Wireless Power Transfer Operated by Time-Modulated Arrays via a Two-Step Procedure

    Directory of Open Access Journals (Sweden)

    Diego Masotti

    2015-01-01

    Full Text Available The paper introduces a novel method for agile and precise wireless power transmission operated by a time-modulated array. The unique, almost real-time reconfiguration capability of these arrays is fully exploited by a two-step procedure: first, a two-element time-modulated subarray is used for localization of tagged sensors to be energized; the entire 16-element TMA then provides the power to the detected tags, by exploiting the fundamental and first-sideband harmonic radiation. An investigation on the best array architecture is carried out, showing the importance of the adopted nonlinear/full-wave computer-aided-design platform. Very promising simulated energy transfer performance of the entire nonlinear radiating system is demonstrated.

  2. Kali: The framework for fine calibration of the LHCb Electromagnetic Calorimeter

    International Nuclear Information System (INIS)

    Belyaev, Ivan; Savrina, Daria; Graciani, Ricardo; Puig, Albert

    2011-01-01

    The precise calibration (at a level of below 1%) of the electromagnetic calorimeter (ECAL) of the LHCb experiment is an essential task for the fulfilment of the LHCb physics program. The final step of this task is performed with two calibration methods using the real data from the experimental setup. It is a very CPU-consuming procedure as both methods require processing of O(10 8 ) events which must be selected, reconstructed and analyzed. In this document we present the Kali framework developed within the LHCb software framework, which implements these two final calibration methods. It is integrated with Grid middleware and makes use of parallelism tools, such as python parallel processing modules, to provide an efficient way, both time and disk wise, for the final ECAL calibration. The results of the fine calibration with the very first data collected by the LHCb experiment will also be presented. With the use of the Kali framework it took only two days of processing and allowed to achieve a calibration accuracy of 2-2.5% for the different ECAL areas.

  3. Modeling and Calibration of a Novel One-Mirror Galvanometric Laser Scanner

    Directory of Open Access Journals (Sweden)

    Chengyi Yu

    2017-01-01

    Full Text Available A laser stripe sensor has limited application when a point cloud of geometric samples on the surface of the object needs to be collected, so a galvanometric laser scanner is designed by using a one-mirror galvanometer element as its mechanical device to drive the laser stripe to sweep along the object. A novel mathematical model is derived for the proposed galvanometer laser scanner without any position assumptions and then a model-driven calibration procedure is proposed. Compared with available model-driven approaches, the influence of machining and assembly errors is considered in the proposed model. Meanwhile, a plane-constraint-based approach is proposed to extract a large number of calibration points effectively and accurately to calibrate the galvanometric laser scanner. Repeatability and accuracy of the galvanometric laser scanner are evaluated on the automobile production line to verify the efficiency and accuracy of the proposed calibration method. Experimental results show that the proposed calibration approach yields similar measurement performance compared with a look-up table calibration method.

  4. Effects of gypsum and bulk density on neutron probe calibration curves

    International Nuclear Information System (INIS)

    Arslan, Awadis; Razzouk, A.K.

    1993-10-01

    The effects of gypsum and bulk density on the neutron probe calibration curve were studied in the laboratory and in the field. The effect of bulk density was negligible for the soil studied in the laboratory, while it was significant for the field calibration. An increase in the slope of moisture content on a volume basis vs. count ratio with increasing gypsum content at the soil was observed in the laboratory calibration. A simple method for correction of the calibration curve for gypsum content was adopted to obtain a specific curve for each layer. The adapted method requires the gypsum fraction to be estimated for each layer and then incorporated in the calibration curve to improve the coefficient of determination. A field calibration showed an improvement of the determination coefficient by introducing bulk density and gypsum fraction, in addition to count ratio using moisture content on a volume basis as a dependent variable in multi linear regression analysis. The same procedure was successful with variable gravel fractions. (author). 18 refs., 3 figs., 2 tabs

  5. Calibration method for projector-camera-based telecentric fringe projection profilometry system.

    Science.gov (United States)

    Liu, Haibo; Lin, Huijing; Yao, Linshen

    2017-12-11

    By combining a fringe projection setup with a telecentric lens, a fringe pattern could be projected and imaged within a small area, making it possible to measure the three-dimensional (3D) surfaces of micro-components. This paper focuses on the flexible calibration of the fringe projection profilometry (FPP) system using a telecentric lens. An analytical telecentric projector-camera calibration model is introduced, in which the rig structure parameters remain invariant for all views, and the 3D calibration target can be located on the projector image plane with sub-pixel precision. Based on the presented calibration model, a two-step calibration procedure is proposed. First, the initial parameters, e.g., the projector-camera rig, projector intrinsic matrix, and coordinates of the control points of a 3D calibration target, are estimated using the affine camera factorization calibration method. Second, a bundle adjustment algorithm with various simultaneous views is applied to refine the calibrated parameters, especially the rig structure parameters and coordinates of the control points forth 3D target. Because the control points are determined during the calibration, there is no need for an accurate 3D reference target, whose is costly and extremely difficult to fabricate, particularly for tiny objects used to calibrate the telecentric FPP system. Real experiments were performed to validate the performance of the proposed calibration method. The test results showed that the proposed approach is very accurate and reliable.

  6. Calibration of robot tool centre point using camera-based system

    Directory of Open Access Journals (Sweden)

    Gordić Zaviša

    2016-01-01

    Full Text Available Robot Tool Centre Point (TCP calibration problem is of great importance for a number of industrial applications, and it is well known both in theory and in practice. Although various techniques have been proposed for solving this problem, they mostly require tool jogging or long processing time, both of which affect process performance by extending cycle time. This paper presents an innovative way of TCP calibration using a set of two cameras. The robot tool is placed in an area where images in two orthogonal planes are acquired using cameras. Using robust pattern recognition, even deformed tool can be identified on images, and information about its current position and orientation forwarded to control unit for calibration. Compared to other techniques, test results show significant reduction in procedure complexity and calibration time. These improvements enable more frequent TCP checking and recalibration during production, thus improving the product quality.

  7. Radioactive contamination in monitors received for calibration

    International Nuclear Information System (INIS)

    Dias, Paulo S.; Santos, Gilvan C. dos; Brunelo, Maria Antonieta G.; Paula, Tiago C. de; Pires, Marina A.; Borges, Jose C.

    2013-01-01

    The Calibration Laboratory - LABCAL, from the Research Center for Metrology and Testing - METROBRAS, MRA Comercio de Instrumentos Eletronicos Ltda., began activities in October 2008 and, in August 2009, decided to establish a procedure for monitoring tests, external and internal, of all packages received from customers, containing instruments for calibration. The aim was to investigate possible contamination radioactive on these instruments. On July 2011, this procedure was extended to packagings of personal thermoluminescent dosemeters - TLD, received by the newly created Laboratory Laboratorio de Dosimetria Pessoal - LDP . In the monitoring procedure were used monitors with external probe, type pancake, MRA brand, models GP - 500 and MIR 7028. During the 37 months in which this investigation was conducted, were detected 42 cases of radioactive contamination, with the following characteristics: 1) just one case was personal dosimeter, TLD type; 2) just one case was not from a packing from nuclear medicine service - was from a mining company; 3) contamination occurred on packs and instruments, located and/or widespread; 4) contamination values ranged from slightly above the level of background radiation to about a thousand fold. Although METROBRAS has facilities for decontamination, in most cases, especially those of higher contamination, the procedure followed was to store the contaminated material in a room used for storage of radioactive sources. Periodically, each package and/or instrument was monitored, being released when the radiation level matched the background radiation. Every contamination detected, the client and/or owner of the instrument was informed. The Brazilian National Energy Commission - CNEN, was informed, during your public consultation for reviewing the standard for nuclear medicine services, held in mid-2012, having received from METROBRAS the statistical data available at the time. The high frequency of contamination detected and the high

  8. Recent developments in the specification and achievement of realistic neutron calibration fields

    International Nuclear Information System (INIS)

    Chartier, J.L.; Kluges, H.; Wiegel, B.; Schraube, H.

    1997-01-01

    In order to calibrate more accurately the neutron dosemeters involved in radiation protection, the concept of 'Realistic Neutron Calibration Fields' is considered as an appropriate alternative solution, making necessary new irradiation facilities which generate well-characterised neutron fields with energy and angular distribution replicating more closely practical workplace conditions. Several experienced laboratories have collaborated on a European project and proposed various approaches which are reviewed in this paper. A short description of the facilities currently in operation is given as well as a few characteristics of the available radiation fields. This description of the state of art is followed by a discussion of the problems to be solved for using such facilities for calibration purposes according to well-specified calibration procedures. (author)

  9. The Grover energy transfer algorithm for relativistic speeds

    Energy Technology Data Exchange (ETDEWEB)

    Garcia-Escartin, Juan Carlos; Chamorro-Posada, Pedro, E-mail: juagar@yllera.tel.uva.e [Dpto. de TeorIa de la Senal y Comunicaciones e Ingenieria Telematica, Universidad de Valladolid, ETSI de Telecomunicacion, Campus Miguel Delibes, Paseo Belen 15, 47011 Valladolid (Spain)

    2010-11-12

    Grover's algorithm for quantum search can also be applied to classical energy transfer. The procedure takes a system in which the total energy is equally distributed among N subsystems and transfers most of it to one marked subsystem. We show that in a relativistic setting the efficiency of this procedure can be improved. We will consider the transfer of relativistic kinetic energy in a series of elastic collisions. In this case, the number of steps of the energy transfer procedure approaches 1 as the initial velocities of the objects become closer to the speed of light. This is a consequence of introducing nonlinearities in the procedure. However, the maximum attainable transfer will depend on the particular combination of speed and number of objects. In the procedure, we will use N elements, as in the classical non-relativistic case, instead of the log{sub 2}(N) states of the quantum algorithm.

  10. A calibration method for fringe reflection technique based on the analytical phase-slope description

    Science.gov (United States)

    Wu, Yuxiang; Yue, Huimin; Pan, Zhipeng; Liu, Yong

    2018-05-01

    The fringe reflection technique (FRT) has been one of the most popular methods to measure the shape of specular surface these years. The existing system calibration methods of FRT usually contain two parts, which are camera calibration and geometric calibration. In geometric calibration, the liquid crystal display (LCD) screen position calibration is one of the most difficult steps among all the calibration procedures, and its accuracy is affected by the factors such as the imaging aberration, the plane mirror flatness, and LCD screen pixel size accuracy. In this paper, based on the deduction of FRT analytical phase-slope description, we present a novel calibration method with no requirement to calibrate the position of LCD screen. On the other hand, the system can be arbitrarily arranged, and the imaging system can either be telecentric or non-telecentric. In our experiment of measuring the 5000mm radius sphere mirror, the proposed calibration method achieves 2.5 times smaller measurement error than the geometric calibration method. In the wafer surface measuring experiment, the measurement result with the proposed calibration method is closer to the interferometer result than the geometric calibration method.

  11. SU-F-E-19: A Novel Method for TrueBeam Jaw Calibration

    Energy Technology Data Exchange (ETDEWEB)

    Corns, R; Zhao, Y; Huang, V [Fraser Valley Cancer Centre - BC Cancer Agency, Surrey, BC (United Kingdom)

    2016-06-15

    Purpose: A simple jaw calibration method is proposed for Varian TrueBeam using an EPID-Encoder combination that gives accurate fields sizes and a homogeneous junction dose. This benefits clinical applications such as mono-isocentric half-beam block breast cancer or head and neck cancer treatment with junction/field matching. Methods: We use EPID imager with pixel size 0.392 mm × 0.392 mm to determine the radiation jaw position as measured from radio-opaque markers aligned with the crosshair. We acquire two images with different symmetric field sizes and record each individual jaw encoder values. A linear relationship between each jaw’s position and its encoder value is established, from which we predict the encoder values that produce the jaw positions required by TrueBeam’s calibration procedure. During TrueBeam’s jaw calibration procedure, we move the jaw with the pendant to set the jaw into position using the predicted encoder value. The overall accuracy is under 0.1 mm. Results: Our in-house software analyses images and provides sub-pixel accuracy to determine field centre and radiation edges (50% dose of the profile). We verified the TrueBeam encoder provides a reliable linear relationship for each individual jaw position (R{sup 2}>0.9999) from which the encoder values necessary to set jaw calibration points (1 cm and 19 cm) are predicted. Junction matching dose inhomogeneities were improved from >±20% to <±6% using this new calibration protocol. However, one technical challenge exists for junction matching, if the collimator walkout is large. Conclusion: Our new TrueBeam jaw calibration method can systematically calibrate the jaws to crosshair within sub-pixel accuracy and provides both good junction doses and field sizes. This method does not compensate for a larger collimator walkout, but can be used as the underlying foundation for addressing the walkout issue.

  12. CP-50 calibration well facility: radiological safety assessment document

    International Nuclear Information System (INIS)

    Orcutt, J.A.; Hill, R.L.

    1984-03-01

    Design features, systems controls, and procedures used in the opeation of the calibration well facility are presented. Site and facility characteristics, as well as routine and nonroutine operations are discussed. Hypothetical incidents and accidents, source control systems, and radiation monitoring considerations are described. 8 references, 35 figures

  13. Project of an integrated calibration laboratory of instruments at IPEN

    International Nuclear Information System (INIS)

    Barros, Gustavo Adolfo San Jose

    2009-01-01

    The Calibration Laboratory of Instruments of Instituto de Pesquisas Energeticas e Nucleares offers calibration services of radiation detectors used in radioprotection, diagnostic radiology and radiotherapy, for IPEN and for external facilities (public and private). One part of its facilities is located in the main building, along with other laboratories and study rooms, and another part in an isolated building called Bunker. For the optimization, modernization and specially the safety, the laboratories in the main building shall be transferred to an isolated place. In this work, a project of an integrated laboratory for calibration of instruments was developed, and it will be an expansion of the current Calibration Laboratory of Instruments of IPEN. Therefore, a series of radiometric monitoring of the chosen localization of the future laboratory was realized, and all staff needs (dimensions and disposition of the study rooms and laboratories) were defined. In this project, the laboratories with X ray equipment, alpha and beta radiation sources were located at an isolated part of the building, and the wall shielding was determined, depending on the use of each laboratory. (author)

  14. Intercomparison programme of dose calibration used in nuclear medicine center in Malaysia

    International Nuclear Information System (INIS)

    Norhayati Abdullah; Abdul Aziz Mohd Ramli; Muhammad Jamal Md Isa; Siti Sara Deraman; Shahrul Azlan Azizan; Nor Azlin Azraai; Md Khairusalih Md Zin

    2010-01-01

    Calibration of dose calibrator is significant in order to ensure that the equipment operates optimally and provides accurate and reliable measurements of the total activity of radiopharmaceuticals before being administered into the patients. Through this work, the response between the secondary standard dose calibrator and users radioactivity measurement are obtained by using standard sources such as 57 Co, 133 Ba, 1 '3 7 Cs and 60 Co. The calibration procedure is in accordance with the NPLs (National Physical Laboratory, United Kingdom) document; Guide No. 93[1] and the IAEA (International Atomic Energy Agency) Technical Report Series No. 454 [2] is used as a reference for maintaining secondary standard dose calibrator. A total of 21 units of dose calibrator from eight nuclear medicine departments comprising five hospitals, two medical centres and one production laboratory were calibrated. The measurement results were inter compared with the national standard equipment and a baseline data was established for future comparison as well as dose optimization purposes. Results showed that the overall response of all dose calibrators are within NPLs tolerance limit of ±10 % except for 5 units which exceed the tolerance limit for radionuclide 133 Ba and 57 Co.(author)

  15. Ray-based calibration for the micro optical metrology system

    Science.gov (United States)

    Yin, Yongkai; Wang, Meng; Li, Ameng; Liu, Xiaoli; Peng, Xiang

    2014-05-01

    Fringe projection 3D microscopy (FP-3DM) plays an important role in micro-machining and micro-fabrication. FP-3DM may be realized with quite different arrangements and principles, which make people confused to select an appropriate one for their specific application. This paper introduces the ray-based general imaging model to describe the FP-3DM, which has the potential to get a unified expression for different system arrangements. Meanwhile the dedicated calibration procedure is also presented to realize quantitative 3D imaging. The validity and accuracy of proposed calibration approach is demonstrated with experiments.

  16. Auto-calibration of Systematic Odometry Errors in Mobile Robots

    DEFF Research Database (Denmark)

    Bak, Martin; Larsen, Thomas Dall; Andersen, Nils Axel

    1999-01-01

    This paper describes the phenomenon of systematic errors in odometry models in mobile robots and looks at various ways of avoiding it by means of auto-calibration. The systematic errors considered are incorrect knowledge of the wheel base and the gains from encoder readings to wheel displacement....... By auto-calibration we mean a standardized procedure which estimates the uncertainties using only on-board equipment such as encoders, an absolute measurement system and filters; no intervention by operator or off-line data processing is necessary. Results are illustrated by a number of simulations...... and experiments on a mobile robot....

  17. Multidetector calibration for mass spectrometers

    International Nuclear Information System (INIS)

    Bayne, C.K.; Donohue, D.L.; Fiedler, R.

    1994-06-01

    The International Atomic Energy Agency's Safeguards Analytical Laboratory has performed calibration experiments to measure the different efficiencies among multi-Faraday detectors for a Finnigan-MAT 261 mass spectrometer. Two types of calibration experiments were performed: (1) peak-shift experiments and (2) peak-jump experiments. For peak-shift experiments, the ion intensities were measured for all isotopes of an element in different Faraday detectors. Repeated measurements were made by shifting the isotopes to various Faraday detectors. Two different peak-shifting schemes were used to measure plutonium (UK Pu5/92138) samples. For peak-jump experiments, ion intensities were measured in a reference Faraday detector for a single isotope and compared with those measured in the other Faraday detectors. Repeated measurements were made by switching back-and-forth between the reference Faraday detector and a selected Faraday detector. This switching procedure is repeated for all Faraday detectors. Peak-jump experiments were performed with replicate measurements of 239 Pu, 187 Re, and 238 U. Detector efficiency factors were estimated for both peak-jump and peak-shift experiments using a flexible calibration model to statistically analyze both types of multidetector calibration experiments. Calculated detector efficiency factors were shown to depend on both the material analyzed and the experimental conditions. A single detector efficiency factor is not recommended for each detector that would be used to correct routine sample analyses. An alternative three-run peak-shift sample analysis should be considered. A statistical analysis of the data from this peak-shift experiment can adjust the isotopic ratio estimates for detector differences due to each sample analysis

  18. Hierarchical calibration and validation framework of bench-scale computational fluid dynamics simulations for solvent-based carbon capture. Part 2: Chemical absorption across a wetted wall column: Original Research Article: Hierarchical calibration and validation framework of bench-scale computational fluid dynamics simulations

    Energy Technology Data Exchange (ETDEWEB)

    Wang, Chao [Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Physical and Computational Sciences Directorate; Xu, Zhijie [Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Physical and Computational Sciences Directorate; Lai, Kevin [Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Physical and Computational Sciences Directorate; Whyatt, Greg [Pacific Northwest National Lab. (PNNL), Richland, WA (United States). Energy and Environment Directorate; Marcy, Peter W. [Los Alamos National Lab. (LANL), Los Alamos, NM (United States); Sun, Xin [Oak Ridge National Lab. (ORNL), Oak Ridge, TN (United States). Energy and Transportation Science Division

    2017-10-24

    Part 1 of this paper presents a numerical model for non-reactive physical mass transfer across a wetted wall column (WWC). In Part 2, we improved the existing computational fluid dynamics (CFD) model to simulate chemical absorption occurring in a WWC as a bench-scale study of solvent-based carbon dioxide (CO2) capture. In this study, to generate data for WWC model validation, CO2 mass transfer across a monoethanolamine (MEA) solvent was first measured on a WWC experimental apparatus. The numerical model developed in this work can account for both chemical absorption and desorption of CO2 in MEA. In addition, the overall mass transfer coefficient predicted using traditional/empirical correlations is conducted and compared with CFD prediction results for both steady and wavy falling films. A Bayesian statistical calibration algorithm is adopted to calibrate the reaction rate constants in chemical absorption/desorption of CO2 across a falling film of MEA. The posterior distributions of the two transport properties, i.e., Henry's constant and gas diffusivity in the non-reacting nitrous oxide (N2O)/MEA system obtained from Part 1 of this study, serves as priors for the calibration of CO2 reaction rate constants after using the N2O/CO2 analogy method. Finally, the calibrated model can be used to predict the CO2 mass transfer in a WWC for a wider range of operating conditions.

  19. Construction of calibration pads facility, Walker Field, Grand Junction, Colorado

    International Nuclear Information System (INIS)

    Ward, D.L.

    1978-08-01

    A gamma-ray spectrometer facility was completed at Walker Field Airport, Grand Junction, Colorado, in November 1976. This report describes spectrometers and their calibration, the construction of the spectrometer facility, the radioelement concentrations, procedures for using the facilites, and environmental considerations

  20. The Geostationary Lightning Mapper: Its Performance and Calibration

    Science.gov (United States)

    Christian, H. J., Jr.

    2015-12-01

    The Geostationary Lightning Mapper (GLM) has been developed to be an operational instrument on the GOES-R series of spacecraft. The GLM is a unique instrument, unlike other meteorological instruments, both in how it operates and in the information content that it provides. Instrumentally, it is an event detector, rather than an imager. While processing almost a billion pixels per second with 14 bits of resolution, the event detection process reduces the required telemetry bandwidth by almost 105, thus keeping the telemetry requirements modest and enabling efficient ground processing that leads to rapid data distribution to operational users. The GLM was designed to detect about 90 percent of the total lightning flashes within its almost hemispherical field of view. Based on laboratory calibration, we expect the on-orbit detection efficiency to be closer to 85%, making it the highest performing, large area coverage total lightning detector. It has a number of unique design features that will enable it have near uniform special resolution over most of its field of view and to operate with minimal impact on performance during solar eclipses. The GLM has no dedicated on-orbit calibration system, thus the ground-based calibration provides the bases for the predicted radiometric performance. A number of problems were encountered during the calibration of Flight Model 1. The issues arouse from GLM design features including its wide field of view, fast lens, the narrow-band interference filters located in both object and collimated space and the fact that the GLM is inherently a event detector yet the calibration procedures required both calibration of images and events. The GLM calibration techniques were based on those developed for the Lightning Imaging Sensor calibration, but there are enough differences between the sensors that the initial GLM calibration suggested that it is significantly more sensitive than its design parameters. The calibration discrepancies have

  1. CALIPSO lidar calibration at 532 nm: version 4 nighttime algorithm

    Directory of Open Access Journals (Sweden)

    J. Kar

    2018-03-01

    Full Text Available Data products from the Cloud-Aerosol Lidar with Orthogonal Polarization (CALIOP on board Cloud-Aerosol Lidar and Infrared Pathfinder Satellite Observations (CALIPSO were recently updated following the implementation of new (version 4 calibration algorithms for all of the Level 1 attenuated backscatter measurements. In this work we present the motivation for and the implementation of the version 4 nighttime 532 nm parallel channel calibration. The nighttime 532 nm calibration is the most fundamental calibration of CALIOP data, since all of CALIOP's other radiometric calibration procedures – i.e., the 532 nm daytime calibration and the 1064 nm calibrations during both nighttime and daytime – depend either directly or indirectly on the 532 nm nighttime calibration. The accuracy of the 532 nm nighttime calibration has been significantly improved by raising the molecular normalization altitude from 30–34 km to the upper possible signal acquisition range of 36–39 km to substantially reduce stratospheric aerosol contamination. Due to the greatly reduced molecular number density and consequently reduced signal-to-noise ratio (SNR at these higher altitudes, the signal is now averaged over a larger number of samples using data from multiple adjacent granules. Additionally, an enhanced strategy for filtering the radiation-induced noise from high-energy particles was adopted. Further, the meteorological model used in the earlier versions has been replaced by the improved Modern-Era Retrospective analysis for Research and Applications, Version 2 (MERRA-2, model. An aerosol scattering ratio of 1.01 ± 0.01 is now explicitly used for the calibration altitude. These modifications lead to globally revised calibration coefficients which are, on average, 2–3 % lower than in previous data releases. Further, the new calibration procedure is shown to eliminate biases at high altitudes that were present in earlier versions and

  2. Calibrating an ellipsometer using x-ray reflectivity

    International Nuclear Information System (INIS)

    Richter, Andrew; Guico, Rodney; Wang, Jin

    2001-01-01

    X-ray reflectivity has been used to find the optical refractive index of polymer thin film in order to calibrate a Stokes ellipsometer for film thickness measurements during the deposition procedure. A thin, spun-cast film of poly(tert-butyl acrylate) (PtBA) was made with a film thickness of ∼500 {angstrom}. An x-ray reflectivity measurement was taken and the data were fit to determine the thickness of the PtBA film and the underlying silicon--oxide layer. This measurement was then used to calculate the optical refractive index for PtBA at the ellipsometer wavelength. Using this value for the refractive index subsequently allowed us to determine the film thickness for a series of PtBA films made by using a number of polymer solution concentrations resulting in film thickness ranging from 100 to 1300 {angstrom}. These film thicknesses were found to be generally the same as those found using x-ray reflectivity. The success of this procedure suggests a useful method for calibrating an ellipsometer for fast in-lab measurements, especially on ultrathin films when simultaneous determination of the film thickness and the refractive index is less reliable

  3. Systematic Calibration for a Backpacked Spherical Photogrammetry Imaging System

    Science.gov (United States)

    Rau, J. Y.; Su, B. W.; Hsiao, K. W.; Jhan, J. P.

    2016-06-01

    A spherical camera can observe the environment for almost 720 degrees' field of view in one shoot, which is useful for augmented reality, environment documentation, or mobile mapping applications. This paper aims to develop a spherical photogrammetry imaging system for the purpose of 3D measurement through a backpacked mobile mapping system (MMS). The used equipment contains a Ladybug-5 spherical camera, a tactical grade positioning and orientation system (POS), i.e. SPAN-CPT, and an odometer, etc. This research aims to directly apply photogrammetric space intersection technique for 3D mapping from a spherical image stereo-pair. For this purpose, several systematic calibration procedures are required, including lens distortion calibration, relative orientation calibration, boresight calibration for direct georeferencing, and spherical image calibration. The lens distortion is serious on the ladybug-5 camera's original 6 images. Meanwhile, for spherical image mosaicking from these original 6 images, we propose the use of their relative orientation and correct their lens distortion at the same time. However, the constructed spherical image still contains systematic error, which will reduce the 3D measurement accuracy. Later for direct georeferencing purpose, we need to establish a ground control field for boresight/lever-arm calibration. Then, we can apply the calibrated parameters to obtain the exterior orientation parameters (EOPs) of all spherical images. In the end, the 3D positioning accuracy after space intersection will be evaluated, including EOPs obtained by structure from motion method.

  4. Improvements to and Comparison of Static Terrestrial LiDAR Self-Calibration Methods

    Directory of Open Access Journals (Sweden)

    Preston Hartzell

    2013-05-01

    Full Text Available Terrestrial laser scanners are sophisticated instruments that operate much like high-speed total stations. It has previously been shown that unmodelled systematic errors can exist in modern terrestrial laser scanners that deteriorate their geometric measurement precision and accuracy. Typically, signalised targets are used in point-based self-calibrations to identify and model the systematic errors. Although this method has proven its effectiveness, a large quantity of signalised targets is required and is therefore labour-intensive and limits its practicality. In recent years, feature-based self-calibration of aerial, mobile terrestrial, and static terrestrial laser scanning systems has been demonstrated. In this paper, the commonalities and differences between point-based and plane-based self-calibration (in terms of model identification and parameter correlation are explored. The results of this research indicate that much of the knowledge from point-based self-calibration can be directly transferred to plane-based calibration and that the two calibration approaches are nearly equivalent. New network configurations, such as the inclusion of tilted scans, were also studied and prove to be an effective means for strengthening the self-calibration solution, and improved recoverability of the horizontal collimation axis error for hybrid scanners, which has always posed a challenge in the past.

  5. Novel real-time alignment and calibration of the LHCb detector in Run2

    CERN Document Server

    AUTHOR|(INSPIRE)INSPIRE-00144085

    2017-01-01

    LHCb has introduced a novel real-time detector alignment and calibration strategy for LHC Run2. Data collected at the start of the fill are processed in a few minutes and used to update the alignment parameters, while the calibration constants are evaluated for each run. This procedure improves the quality of the online reconstruction. For example, the vertex locator is retracted and reinserted for stable beam conditions in each fill to be centred on the primary vertex position in the transverse plane. Consequently its position changes on a fill-by-fill basis. Critically, this new real-time alignment and calibration procedure allows identical constants to be used in the online and offline reconstruction, thus improving the correlation between triggered and offline-selected events. This offers the opportunity to optimise the event selection in the trigger by applying stronger constraints. The required computing time constraints are met thanks to a new dedicated framework using the multi-core farm infrastructur...

  6. Calibration, checking and physical corrections for a new dual-spaced neutron porosity tool

    International Nuclear Information System (INIS)

    Smith, M.P.

    1986-01-01

    A new dual-spaced neutron tool has been developed that features high count rates and improved statistical precision and log repeatability. Environmental corrections including borehole diameter, standoff, and lithology are at acceptable levels for DSN-II. The effects of varying source-to-detector spacings and shielding are summarized. Porosity measurement resolution and statistical precision are discussed and it is indicated how tradeoffs between higher count rates and increased environmental corrections must be considered. The absolute calibration of a standard tool is based on its response to limestone test pits, field data, and theoretical calculations. Test data for actual manufactured tools are presented. Shop calibration and wellsite check procedures are discussed. The advantages of multiposition check operations are explained, including reduced sensitivity to check block positioning and external environment. An analysis is presented of errors from tool manufacturing, calibration, and check procedures. A generalized theory of neutron scattering and absorption has been developed to correct dual-spaced neutron logs for unusual minerals and fluids

  7. Accuracy improvement in a calibration test bench for accelerometers by a vision system

    International Nuclear Information System (INIS)

    D’Emilia, Giulio; Di Gasbarro, David; Gaspari, Antonella; Natale, Emanuela

    2016-01-01

    A procedure is described in this paper for the accuracy improvement of calibration of low-cost accelerometers in a prototype rotary test bench, driven by a brushless servo-motor and operating in a low frequency range of vibrations (0 to 5 Hz). Vibration measurements by a vision system based on a low frequency camera have been carried out, in order to reduce the uncertainty of the real acceleration evaluation at the installation point of the sensor to be calibrated. A preliminary test device has been realized and operated in order to evaluate the metrological performances of the vision system, showing a satisfactory behavior if the uncertainty measurement is taken into account. A combination of suitable settings of the control parameters of the motion control system and of the information gained by the vision system allowed to fit the information about the reference acceleration at the installation point to the needs of the procedure for static and dynamic calibration of three-axis accelerometers.

  8. Accuracy improvement in a calibration test bench for accelerometers by a vision system

    Energy Technology Data Exchange (ETDEWEB)

    D’Emilia, Giulio, E-mail: giulio.demilia@univaq.it; Di Gasbarro, David, E-mail: david.digasbarro@graduate.univaq.it; Gaspari, Antonella, E-mail: antonella.gaspari@graduate.univaq.it; Natale, Emanuela, E-mail: emanuela.natale@univaq.it [University of L’Aquila, Department of Industrial and Information Engineering and Economics (DIIIE), via G. Gronchi, 18, 67100 L’Aquila (Italy)

    2016-06-28

    A procedure is described in this paper for the accuracy improvement of calibration of low-cost accelerometers in a prototype rotary test bench, driven by a brushless servo-motor and operating in a low frequency range of vibrations (0 to 5 Hz). Vibration measurements by a vision system based on a low frequency camera have been carried out, in order to reduce the uncertainty of the real acceleration evaluation at the installation point of the sensor to be calibrated. A preliminary test device has been realized and operated in order to evaluate the metrological performances of the vision system, showing a satisfactory behavior if the uncertainty measurement is taken into account. A combination of suitable settings of the control parameters of the motion control system and of the information gained by the vision system allowed to fit the information about the reference acceleration at the installation point to the needs of the procedure for static and dynamic calibration of three-axis accelerometers.

  9. Calibration service of radiation detectors and dosemeters at IPEN/ Sao Paulo

    Energy Technology Data Exchange (ETDEWEB)

    Potiens, M.P.A.; Caldas, L.V.E. [IPEN, CNEN/SP, Sao Paulo (Brazil)]. e-mail: mppalbu@ipen.br

    2006-07-01

    The Calibration Laboratory of Instituto de Pesquisas Energeticas e Nucleares, IPEN, has already over 25 years been calibrating instruments used in radiation protection and therapy measurements and belonging to hospitals, industries, clinics and other users located in Sao Paulo and in other parts of Brazil. At the present time, the Calibration Laboratory is part of the Radiation Metrology Center and it acts in the Radiation Protection, Radiation Therapy, Nuclear Medicine and Diagnostic Radiology areas, using special set-ups with gamma and beta radiation sealed sources, alpha and beta radiation plane sources and low and intermediate energies of X radiation. Moreover, it has reference instruments for each calibration area with traceability to the Brazilian National Laboratory for Metrology of Ionizing Radiation (secondary standards) and international laboratories (primary standards). The number of tested instruments is increasing annually (from 170 in 1980 to 1871 in 2005), and for the development of new techniques and radiation detectors the continuous improvement of the existing calibration methods is necessary, as well as the establishment of new calibration services to be offered by the Calibration Laboratory for Brazilian and South American users. The objective of this study is to show the evolution of the calibration service developed at IPEN, describing the applied methods and the calibrated instruments types. The quality system implantation process following the basis of the NBR IEC/ISO 17025 standard is also presented with some tools used in the calibration procedures. (Author)

  10. Calibration service of radiation detectors and dosemeters at IPEN/ Sao Paulo

    International Nuclear Information System (INIS)

    Potiens, M.P.A.; Caldas, L.V.E.

    2006-01-01

    The Calibration Laboratory of Instituto de Pesquisas Energeticas e Nucleares, IPEN, has already over 25 years been calibrating instruments used in radiation protection and therapy measurements and belonging to hospitals, industries, clinics and other users located in Sao Paulo and in other parts of Brazil. At the present time, the Calibration Laboratory is part of the Radiation Metrology Center and it acts in the Radiation Protection, Radiation Therapy, Nuclear Medicine and Diagnostic Radiology areas, using special set-ups with gamma and beta radiation sealed sources, alpha and beta radiation plane sources and low and intermediate energies of X radiation. Moreover, it has reference instruments for each calibration area with traceability to the Brazilian National Laboratory for Metrology of Ionizing Radiation (secondary standards) and international laboratories (primary standards). The number of tested instruments is increasing annually (from 170 in 1980 to 1871 in 2005), and for the development of new techniques and radiation detectors the continuous improvement of the existing calibration methods is necessary, as well as the establishment of new calibration services to be offered by the Calibration Laboratory for Brazilian and South American users. The objective of this study is to show the evolution of the calibration service developed at IPEN, describing the applied methods and the calibrated instruments types. The quality system implantation process following the basis of the NBR IEC/ISO 17025 standard is also presented with some tools used in the calibration procedures. (Author)

  11. Calibration of semi-stochastic procedure for simulating high-frequency ground motions

    Science.gov (United States)

    Seyhan, Emel; Stewart, Jonathan P.; Graves, Robert

    2013-01-01

    Broadband ground motion simulation procedures typically utilize physics-based modeling at low frequencies, coupled with semi-stochastic procedures at high frequencies. The high-frequency procedure considered here combines deterministic Fourier amplitude spectra (dependent on source, path, and site models) with random phase. Previous work showed that high-frequency intensity measures from this simulation methodology attenuate faster with distance and have lower intra-event dispersion than in empirical equations. We address these issues by increasing crustal damping (Q) to reduce distance attenuation bias and by introducing random site-to-site variations to Fourier amplitudes using a lognormal standard deviation ranging from 0.45 for Mw  100 km).

  12. Study of the GPS inter-frequency calibration of timing receivers

    Science.gov (United States)

    Defraigne, P.; Huang, W.; Bertrand, B.; Rovera, D.

    2018-02-01

    When calibrating Global Positioning System (GPS) stations dedicated to timing, the hardware delays of P1 and P2, the P(Y)-codes on frequencies L1 and L2, are determined separately. In the international atomic time (TAI) network the GPS stations of the time laboratories are calibrated relatively against reference stations. This paper aims at determining the consistency between the P1 and P2 hardware delays (called dP1 and dP2) of these reference stations, and to look at the stability of the inter-signal hardware delays dP1-dP2 of all the stations in the network. The method consists of determining the dP1-dP2 directly from the GPS pseudorange measurements corrected for the frequency-dependent antenna phase center and the frequency-dependent ionosphere corrections, and then to compare these computed dP1-dP2 to the calibrated values. Our results show that the differences between the computed and calibrated dP1-dP2 are well inside the expected combined uncertainty of the two quantities. Furthermore, the consistency between the calibrated time transfer solution obtained from either single-frequency P1 or dual-frequency P3 for reference laboratories is shown to be about 1.0 ns, well inside the 2.1 ns uB uncertainty of a time transfer link based on GPS P3 or Precise Point Positioning. This demonstrates the good consistency between the P1 and P2 hardware delays of the reference stations used for calibration in the TAI network. The long-term stability of the inter-signal hardware delays is also analysed from the computed dP1-dP2. It is shown that only variations larger than 2 ns can be detected for a particular station, while variations of 200 ps can be detected when differentiating the results between two stations. Finally, we also show that in the differential calibration process as used in the TAI network, using the same antenna phase center or using different positions for L1 and L2 signals gives maximum differences of 200 ps on the hardware delays of the separate

  13. Accurate KAP meter calibration as a prerequisite for optimisation in projection radiography

    International Nuclear Information System (INIS)

    Malusek, A.; Sandborg, M.; Alm Carlsson, G.

    2016-01-01

    Modern X-ray units register the air kerma-area product, PKA, with a built-in KAP meter. Some KAP meters show an energy-dependent bias comparable with the maximum uncertainty articulated by the IEC (25 %), adversely affecting dose-optimisation processes. To correct for the bias, a reference KAP meter calibrated at a standards laboratory and two calibration methods described here can be used to achieve an uncertainty of <7 % as recommended by IAEA. A computational model of the reference KAP meter is used to calculate beam quality correction factors for transfer of the calibration coefficient at the standards laboratory, Q 0 , to any beam quality, Q, in the clinic. Alternatively, beam quality corrections are measured with an energy-independent dosemeter via a reference beam quality in the clinic, Q 1 , to beam quality, Q. Biases up to 35 % of built-in KAP meter readings were noted. Energy-dependent calibration factors are needed for unbiased PKA. Accurate KAP meter calibration as a prerequisite for optimisation in projection radiography. (authors)

  14. Software System for the Calibration of X-Ray Measuring Instruments

    International Nuclear Information System (INIS)

    Gaytan-Gallardo, E.; Tovar-Munoz, V. M.; Cruz-Estrada, P.; Vergara-Martinez, F. J.; Rivero-Gutierrez, T.

    2006-01-01

    A software system that facilities the calibration of X-ray measuring instruments used in medical applications is presented. The Secondary Standard Dosimetry Laboratory (SSDL) of the Nuclear Research National Institute in Mexico (ININ in Spanish), supports activities concerning with ionizing radiations in medical area. One of these activities is the calibration of X-ray measuring instruments, in terms of air kerma or exposure by substitution method in an X-ray beam at a point where the rate has been determined by means of a standard ionization chamber. To automatize this process, a software system has been developed, the calibration system is composed by an X-ray unit, a Dynalizer IIIU X-ray meter by RADCAL, a commercial data acquisition card, the software system and the units to be tested and calibrated. A quality control plan has been applied in the development of the software system, ensuring that quality assurance procedures and standards are being followed

  15. Automation of the Calibration of Reference Dosimeters Used in Radiotherapy

    International Nuclear Information System (INIS)

    Romero Acosta, A.; Gutierrez Lores, S.

    2013-01-01

    Traceability, accuracy and consistency of radiation measurements are essential in radiation dosimetry, particularly in radiotherapy, where the outcome of treatments is highly dependent on the radiation dose delivered to patients. The role of Secondary Standard Dosimetry Laboratories (SSDLs) is crucial in providing traceable calibrations to hospitals, since these laboratories disseminate calibrations at specific radiation qualities appropriate to the use of radiation measuring instruments. These laboratories follow IAEA/WHO guidelines for calibration procedures, often being current and charge measurements described in these guidelines a tedious task. However, these measurements are usually done using modern electrometers which are equipped with a RS-232 interface that allows instrument control from a PC. This paper presents the design and employment of an automated system aimed to the measurements of the radiotherapy dosimeters calibration process for Cobalt-60 gamma rays. A software was developed using Lab View, in order to achieve the acquisition of the charge values measured, calculation of the calibration coefficient and issue of a calibration certificate. A primary data report file is filled and stored in the PC's hard disk. By using this software tool, a better control over the calibration process is achieved, it reduces the need for human intervention and it also reduces the exposure of the laboratory staff. The automated system has been used for the calibration of reference dosimeters used in radiotherapy at the Cuban Secondary Standard Dosimetry Laboratory of the Center for Radiation Protection and Hygiene (Author)

  16. Identifying dominant controls on hydrologic parameter transfer from gauged to ungauged catchments: a comparative hydrology approach

    Science.gov (United States)

    Singh, R.; Archfield, S.A.; Wagener, T.

    2014-01-01

    Daily streamflow information is critical for solving various hydrologic problems, though observations of continuous streamflow for model calibration are available at only a small fraction of the world’s rivers. One approach to estimate daily streamflow at an ungauged location is to transfer rainfall–runoff model parameters calibrated at a gauged (donor) catchment to an ungauged (receiver) catchment of interest. Central to this approach is the selection of a hydrologically similar donor. No single metric or set of metrics of hydrologic similarity have been demonstrated to consistently select a suitable donor catchment. We design an experiment to diagnose the dominant controls on successful hydrologic model parameter transfer. We calibrate a lumped rainfall–runoff model to 83 stream gauges across the United States. All locations are USGS reference gauges with minimal human influence. Parameter sets from the calibrated models are then transferred to each of the other catchments and the performance of the transferred parameters is assessed. This transfer experiment is carried out both at the scale of the entire US and then for six geographic regions. We use classification and regression tree (CART) analysis to determine the relationship between catchment similarity and performance of transferred parameters. Similarity is defined using physical/climatic catchment characteristics, as well as streamflow response characteristics (signatures such as baseflow index and runoff ratio). Across the entire US, successful parameter transfer is governed by similarity in elevation and climate, and high similarity in streamflow signatures. Controls vary for different geographic regions though. Geology followed by drainage, topography and climate constitute the dominant similarity metrics in forested eastern mountains and plateaus, whereas agricultural land use relates most strongly with successful parameter transfer in the humid plains.

  17. Calibration of a rotating accelerometer gravity gradiometer using centrifugal gradients

    Science.gov (United States)

    Yu, Mingbiao; Cai, Tijing

    2018-05-01

    The purpose of this study is to calibrate scale factors and equivalent zero biases of a rotating accelerometer gravity gradiometer (RAGG). We calibrate scale factors by determining the relationship between the centrifugal gradient excitation and RAGG response. Compared with calibration by changing the gravitational gradient excitation, this method does not need test masses and is easier to implement. The equivalent zero biases are superpositions of self-gradients and the intrinsic zero biases of the RAGG. A self-gradient is the gravitational gradient produced by surrounding masses, and it correlates well with the RAGG attitude angle. We propose a self-gradient model that includes self-gradients and the intrinsic zero biases of the RAGG. The self-gradient model is a function of the RAGG attitude, and it includes parameters related to surrounding masses. The calibration of equivalent zero biases determines the parameters of the self-gradient model. We provide detailed procedures and mathematical formulations for calibrating scale factors and parameters in the self-gradient model. A RAGG physical simulation system substitutes for the actual RAGG in the calibration and validation experiments. Four point masses simulate four types of surrounding masses producing self-gradients. Validation experiments show that the self-gradients predicted by the self-gradient model are consistent with those from the outputs of the RAGG physical simulation system, suggesting that the presented calibration method is valid.

  18. Calibration of a Microwave Imaging System Using a Known Scatterer

    DEFF Research Database (Denmark)

    Rubæk, Tonny; Zhurbenko, Vitaliy

    2010-01-01

    the transmission and receiving channels inside of the transceiver modules has been detected. This is most likely caused by the lessthan- perfect isolation of the switches in the modules as well as leakage through the PCB itself. Since the presence of such a leakage signal in the measurements seriously influence...... that the low-amplitude RF signals, available at the terminals of the antennas, only need to travel a very short distance to get to the low-noise amplifier, while the RF as well as the IF signals running to and from the transceiver modules all have significant amplitudes. However, some leakage between...... the imaging capability of the system, it is of interest to remove it. In this work, a calibration procedure capable of removing a constant offset, i.e., the leakage, from the measured signals is presented. The calibration procedure is based on a comparison between the relative change observed between...

  19. Calibration and performance of a stirred benthic chamber

    Science.gov (United States)

    Buchholtz-ten Brink, M. R.; Gust, G.; Chavis, D.

    1989-07-01

    The physical and chemical boundary layer parameters characteristic for a benthic chamber were cross-calibrated by the use of two methods in the laboratory: (1) flush-mounted hot-film sensors, which measure the friction velocity u ∗, and (2) the alabaster dissolution technique, which measures the equivalent film thickness z. Tests of five stirring devices were made, using both techniques, to improve the stirring mechanism in the MANOP Lander flux chambers. The stirring device that was finally implemented consisted of four rods and produced spatially averaged friction velocities u ∗ ranging from 0.1 to 0.5 cm s -1 (i.e. mean film thickness z from 500 to 180 μm) when running at speeds from 3 to 9 rpm. The friction velocity field at the sediment surface is related to the rpm of the stirring device and the penetration depth of the chamber into the sediments; combinations of both can create z and u ∗ inside the chamber that duplicate those of many natural environments. The log-log calibration relationship found between u ∗ and transfer coefficients K' also provides a means to predict the mass-transfer resistance of solutes at the sediment-water interface from measurements of mean bottom stress.

  20. Contributed Review: Absolute spectral radiance calibration of fiber-optic shock-temperature pyrometers using a coiled-coil irradiance standard lamp

    Energy Technology Data Exchange (ETDEWEB)

    Fat’yanov, O. V., E-mail: fatyan1@gps.caltech.edu; Asimow, P. D., E-mail: asimow@gps.caltech.edu [Division of Geological and Planetary Sciences 252-21, California Institute of Technology, Pasadena, California 91125 (United States)

    2015-10-15

    We describe an accurate and precise calibration procedure for multichannel optical pyrometers such as the 6-channel, 3-ns temporal resolution instrument used in the Caltech experimental geophysics laboratory. We begin with a review of calibration sources for shock temperatures in the 3000-30 000 K range. High-power, coiled tungsten halogen standards of spectral irradiance appear to be the only practical alternative to NIST-traceable tungsten ribbon lamps, which are no longer available with large enough calibrated area. However, non-uniform radiance complicates the use of such coiled lamps for reliable and reproducible calibration of pyrometers that employ imaging or relay optics. Careful analysis of documented methods of shock pyrometer calibration to coiled irradiance standard lamps shows that only one technique, not directly applicable in our case, is free of major radiometric errors. We provide a detailed description of the modified Caltech pyrometer instrument and a procedure for its absolute spectral radiance calibration, accurate to ±5%. We employ a designated central area of a 0.7× demagnified image of a coiled-coil tungsten halogen lamp filament, cross-calibrated against a NIST-traceable tungsten ribbon lamp. We give the results of the cross-calibration along with descriptions of the optical arrangement, data acquisition, and processing. We describe a procedure to characterize the difference between the static and dynamic response of amplified photodetectors, allowing time-dependent photodiode correction factors for spectral radiance histories from shock experiments. We validate correct operation of the modified Caltech pyrometer with actual shock temperature experiments on single-crystal NaCl and MgO and obtain very good agreement with the literature data for these substances. We conclude with a summary of the most essential requirements for error-free calibration of a fiber-optic shock-temperature pyrometer using a high-power coiled tungsten halogen