Sample records for calibration calibracion del

  1. IIE`s wind tunnel calibration; Calibracion del tunel de viento del IIE

    Pena Garcia, Raymundo [Instituto de Investigaciones Electricas, Cuernavaca (Mexico)


    The calibration of a wind tunnel is performed in such a way as to warrant a very low turbulence grade. When there is recently built tunnel, as is the case of the IIE`s tunnel, the turbulence in its testing chambers is large; for this reason it is necessary to integrate in it aerodynamic devices and elements capable of reducing it. At the end of the calibration studies can be performed in models with controlled scale. From these and from the results obtained it will be decided if the designed prototypes are built or modified. [Espanol] La calibracion de un tunel de viento se realiza de tal forma que garantiza un grado de turbulencia muy bajo. Cuando se tiene un tunel recien construido, como es el caso del tunel de viento del IIE, la turbulencia en sus camaras de prueba es grande; por lo que es necesario integrarle dispositivos y elementos aerodinamicos que sean capaces de reducirla. Al terminar la calibracion pueden realizarse estudios en modelos con escala controlada. De estos y de los resultados que se obtengan se decidira si se construyen o se modifican los prototipos disenados.

  2. Automation of dosimeters calibration for radiotherapy in secondary dosimetric calibration laboratory of the CPHR; Automatizacion de la calibracion de dosimetros de radioterapia en el laboratorio secundario de calibracion dosimetrica del CPHR

    Acosta, Andy L. Romero; Lores, Stefan Gutierrez, E-mail: [Centro de Proteccion e Higiene de las Radiaciones (CPHR), La Habana (Cuba)


    This paper presents the design and implementation of an automated system for measurements in the calibration of reference radiation dosimeters. It was made a software application that performs the acquisition of the measured values of electric charge, calculated calibration coefficient and automates the calibration certificate issuance. These values are stored in a log file on a PC. The use of the application improves control over the calibration process, helps to humanize the work and reduces personnel exposure. The tool developed has been applied to the calibration of dosimeters radiation patterns in the LSCD of the Centro de Proteccion e Higiene de las Radiaciones, Cuba.

  3. Calibration of the indicators of the product KERMa-AREA in radiodiagnostic equipment; Calibracion de los indicadores del producto KERMA-AREA en los equipos de radiodiagnostico

    Ginjaume, M.; Jarvinen, H.; Turak, O.


    Currently the IAEA proposes two procedures for the calibration of the determination of the PKA systems integrated in the x-ray equipment. The traditional method is to measure air kerma at a certain distance from the focus through an ionization chamber reference and then the area of the field of radiation in that position. The alternative method is based on measuring directly the product kerma-area using equipment calibrated in this magnitude. In both cases the calibration procedure takes into account the position of the tube on the stretcher of the patient. The objective of this study was to compare the practical application of both calibration procedures as well as assess the accuracy of indication of the PKA in the consoles of the radiology equipment. (Author)

  4. Calibration and start-up of the system of radiotherapy extract image-guided; Calibracion y puesta en marcha del sistema de radioterpia guiada por la imagen exactrac

    Clemente Gutierrez, I.; Perez Vara, C.; Prieto Villacorta, M.


    The accuracy in the administration of external radiotherapy treatments may be increased through the use of guidance systems by image. Constitute a particular case x-rays teams independent of the treatment unit. Among them is the ExacTrac (Brainlab) system. The objective of this work is to briefly introduce the procedure followed in the calibration and implementation of such a system. (Author)

  5. Dosimetry through the Secondary Laboratory of Dosimetric Calibration of Mexico; Dosimetria a traves del Laboratorio Secundario de Calibracion Dosimetrica de Mexico

    Tovar M, V.M.; Alvarez R, J.T.; Medina O, V.P.; Vergara M, F.; Anaya M, R.; Cejudo A, J.; Salinas L, B. [ININ, A.P. 18-1027, 11801 Mexico D.F. (Mexico)


    In the beginnings of the sixty years an urgent necessity is presented mainly in the developing countries, of improving in important form the accuracy in the dosimetry of external faces in therapy of radiations (radiotherapy centers), mainly in the calibration of ''clinical dosemeters''. In 1976 the International Atomic Energy Agency, (IAEA), and the World Health Organization, (WHO), they carried out a mutual agreement with regard to the establishment and operation of a net of Secondary Patron Laboratories of Dosimetry, (LSCD). The necessity to establish measure patterns in the field of the dosimetry of the ionizing radiations, is necessary, to have an accuracy but high in the dosimetry of the radiation beams in therapy which is highly dependent of the dose given to the tumor of those patient with cancer. Similar levels of accuracy are required in protection measures to the radiation with an acceptable smaller accuracy, however, when the personal dosemeters are used to determine the doses received by the individuals under work conditions, such mensurations in therapy of radiations and radiological protection will have traceability through a chain of comparisons to primary or national patterns. The traceability is necessary to assure the accuracy and acceptability of the dosimetric measures, as well as, the legal and economic implications. The traceability is also necessary in the dosimetry of high dose like in the sterilization of different products. The main function of the LSCD is to provide a service in metrology of ionizing radiations, maintaining the secondary or national patterns, which have a traceability to the International System of measures, which is based for if same in the comparison of patterns in the Primary Laboratories of Dosimetry (LPD) under the auspice of the International Office of Weights and Measure (BIPM). The secondary and national patterns in the LSCD constitute in Mexico, the national patterns of the magnitudes in the

  6. Characterization of the neutron field of the {sup 241}AmBe in a calibration room; Caracterizacion del campo de neutrones del {sup 241} AmBe en una sala para calibracion

    Vega C, H.R. [UAZ, A.P. 336, 98000 Zacatecas (Mexico); Gallego, E.; Lorente, A. [Universidad Politecnica de Madrid, C. Jose Gutierrez Abascal 2, 28006 Madrid (Spain)] e-mail:


    The field of neutrons produced by an isotopic source of neutrons of {sup 241} Am Be had been characterized. The characterization was carried out modeling those relevant details of the calibration room and simulating the neutron transport at different distances of the source. The calculated spectra were used to determine the equivalent environmental dose rate. A series of experiments were carried out with the Bonner sphere spectrometric system to measure the spectra in the same points where the calculations were carried out and with these spectra the rates of environmental dose were calculated. By means of a one sphere dosemeter type Berthold the rates of environmental dose were measured. To the one to compare the calculated spectra and measured its were found small differences in the group of the thermal neutrons due to the elementary composition used during the simulation. When comparing the derived rates starting from the calculated spectra with those measured it was found a maxim difference smaller to 13%. (Author)

  7. Parameters of calibration of the measurement system of {sup 222} Rn based in LR-115; Parametros de calibracion del sistema de medicion de {sup 222} Rn basado en LR-115

    Garcia, M.L.; Mireles, F.; Quirino, L.; Davila, I.; Lugo, F.; Pinedo, J.L. [CREN-UAZ, 98068 Zacatecas (Mexico); Chavez, A. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)]. e-mail:


    Since the SSNTD technique (Solid State Nuclear Track Detection) it was discovered it has been used as passive method for the detection of subnuclear particles in great variety of fields of the science. The use of the technique in measurements of {sup 222} Rn in air have already been established implying better methodologies in the exhibition to the environment until their engraving and reading processes. The SSNTD technique is since a method by comparison since the material it can be used a single time, therefore it requires of calibration in one controlled radon atmosphere, using gauged standards. The objective of this work is to show the calibration of the devices used as radon monitors based on SSNTD. The material used as SSNTD is LR-115 Il. The standardization of the parameters used in the exhibition to radon in air, engraving and reading process, its are based on the response of the LR-115 Il, the one arrangement of the device, engraving speed and mainly the calibration factor. They are considered two types of monitors: Open camera and Closed camera, the difference among the calibration factors of both cameras is the percentage of the descendants of radon in the open camera. The standardized parameters are operation voltage of the counting system; temperature, time and concentration of the engraving solution; and thickness. (Author)

  8. Description of the XXXIV ARCAL Project {sup R}epairing and calibration of electrometers and ionization chambers used in radiotherapy; Descripcion del proyecto ARCAL XXXIV {sup R}eparacion y calibracion de electrometros y camaras de ionizacion utilizados en radioterapia

    Cruz E, P.; Villaverde L, A. [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, 11801 Mexico D.F. (Mexico)


    The technological tools from what the humanity has for the illnesses diagnosis and the cancer treatment, are based in great extent in the use of ionizing radiations. This situation worries to the International Atomic Energy Agency (IAEA), which has implemented technical cooperation programs for protecting the human health. In Latin America the ARCAL program (Regional Agreement of Cooperation for Promotion of Nuclear Science and Technology in Latina America and the Caribbean was created. The Project ARCAL XXXIV has as objective to establish three regional centers of repairing, maintenance and electric calibration of clinical dosemeters, equipment made up for an ionization chamber and an electrometer which is used in radiotherapy to generate calibration procedures, personnel training, establishment of an intercomparison net for the electrometers control used as standards and designing current intensity sources which serve as work standards for each one of the participant countries, Mexico is one of them. (Author)

  9. Calibration of the identiFINDER detector for the iodine measurement in thyroid using the Monte Carlo method; Calibracion del detector identiFINDER para la medicion de yodo en tiroides utilizando el metodo Monte Carlo

    Ramos M, D.; Yera S, Y.; Lopez B, G. M.; Acosta R, N.; Vergara G, A., E-mail: [Centro de Proteccion e Higiene de las Radiaciones, Calle 20 No. 4113 e/ 41 y 47, Playa, 10600 La Habana (Cuba)


    This work is based on the determination of the detection efficiency of {sup 125}I and {sup 131}I in thyroid of the identiFINDER detector using the Monte Carlo method. The suitability of the calibration method is analyzed, when comparing the results of the direct Monte Carlo method with the corrected, choosing the latter because the differences with the real efficiency stayed below 10%. To simulate the detector their geometric parameters were optimized using a tomographic study, what allowed the uncertainties minimization of the estimates. Finally were obtained the simulations of the detector geometry-point source to find the correction factors to 5 cm, 15 cm and 25 cm, and those corresponding to the detector-simulator arrangement for the method validation and final calculation of the efficiency, demonstrating that in the Monte Carlo method implementation if simulates at a greater distance than the used in the Laboratory measurements an efficiency overestimation can be obtained, while if simulates at a shorter distance this will be underestimated, so should be simulated at the same distance to which will be measured in the reality. Also, is achieved the obtaining of the efficiency curves and minimum detectable activity for the measurement of {sup 131}I and {sup 125}I. In general is achieved the implementation of the Monte Carlo methodology for the identiFINDER calibration with the purpose of estimating the measured activity of iodine in thyroid. This method represents an ideal way to replace the lack of patterns solutions and simulators assuring the capacities of the Internal Contamination Laboratory of the Centro de Proteccion e Higiene de las Radiaciones are always calibrated for the iodine measurement in thyroid. (author)

  10. The CIEMAT Neutron Standards Laboratory (LPN): National Reference Facility for the Calibration of Neutron Measurement Equipment; El Laboratorio de Patrones Neutronicos del CIEMAT: instalacion de referencia nacional para la calibracion de equipos de medida neutronica

    Mendez, R.


    This is a new facility that, in keeping with the recommendations of standard ISO 8529:1, has 20m5-Be and 252Cf neutron sources stored in water inside a 9 m x 7.5 m x 8 m Irradiation Room with 1.25 m concrete walls that is accessed through a 6.5 ton sliding shielded door. These sources are remotely handled from the Control Room by means of a system that places them in the geometric center of the room and aligns them with the equipment or materials to be irradiated arranged on a base plate. The facility has a series of interlocks that guarantee safety during operations with sources, and the verification measurements have demonstrated the suitability of the shielding used. The LPN will be the answer to the demand for a national laboratory for neutron measurement equipment calibration and equipment and material irradiation under controlled conditions, and it will position our country at the same level as our European partners. (Author)

  11. Control console of the gamma calibration room; Consola de control de la Sala de Calibracion Gamma

    Vilchis P, A.E.; Romero G, M. [Instituto Nacional de Investigaciones Nucleares, Ingenieria Electronica, A.P. 18-1027, 11801 Mexico D.F. (Mexico)


    The Nuclear Centre of Mexico has a Ionizing Radiation Metrology Center (CMRI). This is in charge of the calibration in Mexico and Latin America of equipment dedicated to radiation measurement as industrial, medical as other fields. The importance to ensure that the equipment stay justly calibrated, it is imposed the necessity of automating the different rooms which the CMRI has. in this case it will be exposed the Calibration room for gamma radiation type. The operation of this application was carried out with the LabVIEW development platform and also in C language. The hardware associated is: personal computer with two cards using the 8255 device, 16 channels with optical isolation to manage input/output TTL type, 16 channels with optical isolation to management of charges to 127 V a.c., and 2 channels for 90V d.c. motors. (Author)

  12. Calibration curves for biological dosimetry; Curvas de calibracion para dosimetria biologica

    Guerrero C, C.; Brena V, M. [ININ, A.P. 18-1027, 11801 Mexico D.F. (Mexico)]. E-mail


    The generated information by the investigations in different laboratories of the world, included the ININ, in which settles down that certain class of chromosomal leisure it increases in function of the dose and radiation type, has given by result the obtaining of calibrated curves that are applied in the well-known technique as biological dosimetry. In this work is presented a summary of the work made in the laboratory that includes the calibrated curves for gamma radiation of {sup 60} Cobalt and X rays of 250 k Vp, examples of presumed exposure to ionizing radiation, resolved by means of aberration analysis and the corresponding dose estimate through the equations of the respective curves and finally a comparison among the dose calculations in those people affected by the accident of Ciudad Juarez, carried out by the group of Oak Ridge, USA and those obtained in this laboratory. (Author)

  13. Use of the electronic system of portal image for quality control and calibration of the position of blades of multi leaf collimation system of an accelerator Elekta synergy; Uso del sistema electronico de imagen portal para el control de calidad y la calibracion de la posicion de las laminas del sistema de colimacion multilaminas de un acelerador elekta synergy

    Sanchez Merino, G.; Cortes Rodicio, J.; Lope Lope, R.; Martin gonzalez, T.; Garcia Fidalgo, M. A.


    We present our methodology for quality control of the positioning of the blades using electronic portal imaging system (ILD). Also an example of its usefulness for calibration multi leaf collimation system (MLC). (Author)

  14. Implementation of the method air-kerma product area in KAP camera calibration with reference qualities of X-ray series at the SSDL RQR of Mexico; Implementacion del metodo producto kerma en aire-area en la calibracion de camaras KAP con calidades de referencia de rayos X de la serie RQR en el LSCD de Mexico

    Cejudo, Jesus; Tovar, Victor M., E-mail:, E-mail: [lnstituto Nacional de lnvestigaciones Nucleares (DMRI/LSDC/lNlN), Ocoyoacac (Mexico). Centro Nuclear Dr. Nabor Carrillo Flores. Laboratorio Secundario de Calibracion Dosimetrica


    The X-Ray machines, at a reference laboratory for the instrument calibration in diagnostic radiology, should compliance with the ISO requirements. Sometimes there is not available as much laboratories as needed in Latin American countries. So this project shows the KAP ionization chamber implementation method using the TRS-457 radiation quality from the IAEA at the SSDL of Mexico. The KAP instruments calibration method consists in doing a substitution comparison using a standard reference with traceability to a primary laboratory and a transmission-monitoring chamber that measures the number of photons of the X-ray primary beam. A KAP chamber calibration requires a special array that consists in collocating the chamber in two different positions of its calibration process. Then, with air kerma-area product coefficient together with a corrected electrometer measure at referential conditions, the patient dosimetry magnitudes are calculated. The dosimetry necessity at hospitals always will be in function of possessing a highly reliable calibration coefficient chamber for making these measures. That dosimetry results will help in reducing the total or partial irradiation emitted to the human body of the patient. This is how stochastic risks will be lessened due to diagnostic studies. The purpose of this project is to have a synergy with calibration for making known that the SSDL of Mexico has the technical capacity to act as a link between primary standard dosimetry laboratories and the ionizing radiation equipment users who require that their KAP chamber have traceability from the primary standard to the user. (author)

  15. Geometric calibration of the system of micro-CBCT; Calibracion geometrica de un sistema micro-CBCT

    Martinez Rodriguez, I.; Chevalier del Rio, M.; Alieva, T.; Garcia Pinto, D.


    The image reconstruction from projections generated with a beam computed tomography system conical (CBCT) requires very precise calibration of a large number of variables that characterize the geometry of the System. Currently, there are several methods of calibration1consistent in the use of specific mannequins. In this work has been used a methodology based on a simple dummy images 2 that has proven results similar or even better compared to other more complex methods. (Author)

  16. Calibration of a PIXE system for analysis of environmental aerosols; Calibracion de un sistema PIXE para analisis de aerosoles ambientales

    Aldape, F.; Diaz, R.V.; Anguiano, G.; Flores M, J. [ININ, 52045 Ocoyoacac, Estado de Mexico (Mexico)


    The results of the calibration of the PIXE system for analysis of environmental aerosols in the CNM is presented. The system was calibrated in order to lead to the project study of environmental aerosols using PIXE from the qualitative dimension to the quantitative one. Part of this work it was carried out in collaboration with the expert of the IAEA Dr. D. Crumpton. The calibration of the system it was achieved bombarding thin targets of well-known composition with protons of 2.5 MeV. Using this technique it was determined the absolute elementary concentration in samples of atmospheric aerosols. As an example the calculation of the lead concentration in one of them whose result was of 662 {+-} 49 ng m{sup -3} is presented. This value agrees with the measurements made in the Crocker Nuclear Lab of the UCD. Its are also discussed the reproducibility and precision of the used technique. (Author)

  17. Calibration of a NE213 detector for neutron spectroscopy; Calibracion de un detector de NE213 para espectroscopia de neutrones

    Blazquez Martinez, J.; Butragueno Casado, J. L.


    This work describes the experimental way followed for getting the calibration of a NE213 detector with a beam of neutrons from the J.E.N. 2 MeV Van de Graaff and using at once pulse shape discrimination. Detector has been used for measuring the spectrum of the fast reactor CORAL-1. There is also included an experimental method in order to get with precision where the Compton edge is placed on the electron spectrum. (Author) 9 refs.

  18. MEPSOCON project: Calibration of Radiometers for High Solar Irradiance; Proyecto MEPSOCON: Calibracion de Radiometros de Alta Irradiancia Solar

    Ballestrin, J.; Rodriguez-Alonso, M.


    The development of central receiver solar plants is a currently emerging field into renewable energies. For several years various receiver prototypes have been evaluated at the Plataforma Solar de Almeria (PSA). The measurement of the incident solar power on the receiver aperture is fundamental to the estimation of its efficiency. Many factors interfere with this measurement and consequently accuracy is very low. This uncertainty is transmitted to the design of the final solar plant and thereby to its price. The sensors used for this measurement are of small size in comparison with the receiver apertures, therefore different systems are necessary to obtain the incident solar power on the receiver aperture from the individual radiometer measurements. This report presents calibration procedures for the sensor used on the measurement of high solar irradiance and the analysis of the different factors affecting the incident power measurement to significantly reduce its uncertainty. (Author) 16 refs.

  19. Calibration curves for quantifying praseodymium by UV-VIS; Curvas de calibracion para cuantificar praseodimio por UV-VIS

    Gonzalez M, R.; Lopez G, H.; Rojas H, A. [ININ, 52750 La Marquesa, Estado de Mexico (Mexico)]. e-mail:


    The UV-Vis spectroscopic technique was used to determine the absorption bands depending on the concentration from the praseodymium solutions at pH3. Those more appropriate were in the wavelength of 215 nm, for concentrations of 0.0001-0.026 M, of 481nm, 468 nm and 443 nm, for concentrations of 0.026-0.325 M, and of 589 nm, for concentrations of 0.026-0.65 M of the praseodymium. To these wavelengths the calibration curves were determined, which presented correlation coefficients between 0.9976 and 0.9999 except of the absorption of 589 nm that gave R{sup 2} = 0.9014. (Author)

  20. Calibration of the hydraulic model of a full-scale activated sludge plant; Calibracion hidraulica a escala real de un reactor de lodos activados

    Fall, Cheikh [Universidad Autonoma del Estado de Mexico (Mexico); Loaiza-Navia, Jimmy [Servicios de Agua y Drenaje de Monterrey (Mexico)


    When planning to simulate a wastewater treatment plant (WWTP) with the activated sludge model number 1 (ASM1), one of the first requirements is to determine the hydraulic model of the reactor. The aim of this study was to evaluate the hydrodynamic regime of the aeration tank of a municipal WWTP by using a rhodamine tracer test and the Aquasim simulation software. A pre-simulation was performed in order to quantify the appropriate colorant mass, set up a sampling plan and evaluate the anticipated visual impact of the tracer test in the river receiving the treated effluents. A tracer test and dynamic flow measurements were carried out, the results of which served to establish and calibrate the hydraulic model. The evaluated tank was physically built as a plug-flow reactor subdivided in 7 compartments, but the study revealed that it is best represented by a model with 5 virtual mixed reactors in series. Through the study, the approach of using a WWTP simulator for hydraulics calibration was shown to be a powerful and flexible tool for designing a tracer test and for identifying adequate tank-in-series models of full-scale activated sludge aeration tanks. [Spanish] Cuando se planea simular una planta de tratamiento con base en el modelo numero 1 de lodos activados (ASM1), uno de los primeros requisitos es determinar el modelo hidraulico del reactor. En este trabajo se estudio el regimen hidrodinamico del tanque de accion de una planta de tratamiento de aguas residuales municipales (PTAR), utilizando una prueba de trazador con rodamina y un programa de simulacion (Aquasim). Se realizo una prueba de trazador con el experimento, lo que permitio determinar la cantidad requerida de trazador, fijar los intervalos de muestreo y limitar el impacto visual anticipado del colorante sobre el rio que recibe el efluente tratado. Se llevaron a cabo la prueba de trazador y la medicion de los perfiles dinamicos de caudales, cuyos resultados sirvieron para establecer y calibrar el

  1. Development of a procedure of calibration of meters of product Kerma-Area; Desarrollo de un procedimiento de calibracion de medidores de producto Kerma-Area

    Ginjaume, M.; Roig, M.; Amores, M.; Ortega, X.


    The aim of this paper is to present the calibration procedure developed, uncertainties associated and scope of accredited degree. Also the objectives of EURAMET 1177 and calibrations are described conducted by our laboratory in this intercomparison. (Author)

  2. Calibration of film radiochromic EBT2 for sources of I-125 encapsulated; Calibracion de pelicula radiocromica EBT2 para fuentes de I-125 encapsulado

    Huerga Cabrerizo, C.; Luquero Llopis, N.; Torre Hernandez, I. de la; Ferrer Garcia, C.; Corredoira silva, E.; Serrada Hierro, A.


    This paper determines the calibration curve in absolute dose for sources of I-125 encapsulated to estimate its uncertainty. In order to assess energy dependence is compared with the obtained for an accelerator of 6MV calibration curve. (Author)

  3. Calibration of a Plastic Classification System with the CCW Model; Calibracion de un Sistema de Clasification de Plasticos segun el Modelo CCW

    Barcala Riveira, J. M.; Fernandez Marron, J. L.; Alberdi Primicia, J.; Navarrete Marin, J. J.; Oller Gonzalez, J. C.


    This document describes the calibration of a plastic Classification system with the CCW model (Classification by Quaternions built Wavelet Coefficients). The method is applied to spectra of plastics usually present in domestic wastes. Obtained results are showed. (Author) 16 refs.

  4. Calibration and quality control of a multi leaf collimator using linear array of detectors; Calibracion y control de calidad de un colimador multilaminas mediante array lineal de detectores

    Suero Rodrigo, M. A.; Marques Fraguela, E.


    The protocol for calibration and quality control established by Siemens for the multi leaf collimator (MLC) of Primus electron linear accelerators, using the light field coincidence with the beam of radiation to determine the position of the blades. In this paper, we illustrate the use of a calibration method for determining the positions of MLC plates radiologically with the help of a linear array of detectors, based on the proposal Lopes et al (2007).

  5. Evaluation of different calibration curves QA of IMRT plans with radiochromic films; Evaluacion de diversas curvas de calibracion QA de planes de IMRT con peliculas radiocromicas

    Hernandez Rodriguez, J.; Martin Rincon, C.; Garcia Repiso, S.; Ramos Paheo, J. A.; Verde Velasco, J. M.; Sena Espinel, E. de


    The non-linear relationship between dose and the optical density, characteristic plates radiochromic Gafchromic EBT and EBT2, has been studied by various authors, whose publications are proposed different functional forms that fit the specific values measured curves that allow the full range of useful dose calibration. The objective of the work focuses on evaluating the influence of the use of different calibration curves in the dose measurement for quality control of IMRT treatments. (Author)

  6. Monte Carlo simulation: tool for the calibration in analytical determination of radionuclides; Simulacion Monte Carlo: herramienta para la calibracion en determinaciones analiticas de radionucleidos

    Gonzalez, Jorge A. Carrazana; Ferrera, Eduardo A. Capote; Gomez, Isis M. Fernandez; Castro, Gloria V. Rodriguez; Ricardo, Niury Martinez, E-mail: [Centro de Proteccion e Higiene de las Radiaciones (CPHR), La Habana (Cuba)


    This work shows how is established the traceability of the analytical determinations using this calibration method. Highlights the advantages offered by Monte Carlo simulation for the application of corrections by differences in chemical composition, density and height of the samples analyzed. Likewise, the results obtained by the LVRA in two exercises organized by the International Agency for Atomic Energy (IAEA) are presented. In these exercises (an intercomparison and a proficiency test) all reported analytical results were obtained based on calibrations in efficiency by Monte Carlo simulation using the DETEFF program.

  7. Transference patron for the calibration of activemeters for use in nuclear medicine; Patron de transferencia para la calibracion de activimetros de uso en medicina nuclear

    Cortes P, A.; Garcia D, O.C.; Ortiz P, I.; Becerril V, A. [Instituto nacional de Investigaciones Nucleares, Laboratorio de Patrones Radiactivos, Departamento de Metrologia, A.P. 18-1027. C.P. 11801 Mexico D.F. (Mexico)


    With the object to improve the calibration service of the activemetersof the users (nuclear medicine centres) so internal as external users at ININ and initiating the establishment of the traceability of the activity measures that carry out in the country, with respect to the National Patron of Nuclear Activity No. ININ-PNM-2, it is initiated the characterization and the performance control of the activemeter make Capintec, model CRC-7BT of the laboratory of Radioactive Patrons (LPR) of the Department of Metrology for to convert it in a transference patron. This characterization and control are based in the sustained review of the activemeter calibration for the I-131, Mo-99 and Sm-153 radioisotopes with activities from 100 mCi until some {mu} Ci and the Tc-99m, In-111, Tl-201, Ga-67, Cs-137, Co-60, and Ba-133 with activities of the range of some {mu} Ci. To verify the good performance of the instrument it was revised the linearity of its scale, the stability of readings, the variation of readings respect of the source position inside the activemeter well and it was determined the calibration factors for each one of the radiosotopes mentionated. The radioactive sources of mid-short life that are used which consist in radioactive solutions of 10 cm{sup 3} contained in polyethylene small bottles of 25 cm{sup 3} elaborated in the LPR and whose activity is measured with respect to the National Patron ININ-PNM-2. The sources of mid-large life are ampoules with 5 cm{sup 3} of radioactive solution, calibrated in activity by the primary laboratory, LNHB (formerly LMRI) of France. (Author)

  8. Energy Calibration of a Silicon Detector Using Pure Beta-Emitters; Calibracion Energetica de un Detector de Silicio Mediante Emisores Beta Puros

    Borras, C.; Arcos, J. M. los


    Energy calibration of SI detectors used in electron spectroscopy 13 commonly performed with conversion electron sources or monoenergetic electrons beams, which are preferred against beta emitters due to the problems arising from their continuous spectra. This paper presents a simple calibration procedure for a PIP-type silicon detector, using 14C, 147Pm, 99{sup T}c and 45Ca sources, that is based on the correspondence between the average channel observed in the experimental spectrum and the mean energy evaluated from the theoretical Fermi distribution for each nuclide. First, a method for evaluating the average channel in the experimental spectrum distorted by the electronic noise is described and its uncertainty estimated. Then, the channel-energy relation ship is established by least squares fitting modified to account for uncertainties in both variables.The calibration has been successfully verified with 147Pm and '09cd sources, showing discrepaneles not greater than 2.5%, within the uncertainties due to the detector resolution and the sources features. (Author)

  9. Characterization of different qualities in X-rays, for instruments calibration in radiological protection; Caracterizacion de diferentes calidades en rayos X, para calibracion de instrumentos en proteccion radiologica

    Cejudo A, J.; Tovar M, V. M.; Vergara M, F., E-mail: jesus.cejudo@inin.gob.m [ININ, Departamento de Metrologia de Radiaciones Ionizantes, Laboratorio Secundario de Calibracion Dosimetrica, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)


    In the Secondary Laboratory of Dosimetric Calibration in Mexico was realized the qualities characterization of the series X-rays RQR reported in the International Code of Practices in Dosimetry and Diagnostic Radiology No. 457, using attenuator filters of high purity aluminum and ionizing radiation equipment of inherent filtration of 4 mm Be with and emergent X-radiation beam of 40 grades. For the attenuation was used a geometric arrangement with three beam limiters and a monitor camera prepared on the established form in the mentioned technical report, and a spherical ionization chamber with collection volume of 3.6 cm{sup 3}, aligning its geometric center with the focus of X-rays tube to get that the incident radiation direction will be perpendicular to the ionization chamber. From the perspective of the radiological protection is important to know the X-radiation quality for the application dedicated to the instruments calibration and can to give to these the traceability to a reference laboratory, this way the quality combination and reference chamber can give as a result a procedure for the evaluation of the entrance in surface dose to estimate the dose orientate levels, specified in the basic standards of safety. (Author)

  10. Calibration of a plastic detector PVT of large dimensions for the detection gamma radiation; Calibracion de un detector plastico de PVT de grandes dimensiones para la deteccion de radiacion gamma

    Armenteros Carmona, J. C.; Suarez navarro, M. J.; Gonzalez-gonzalez, J. A.; Pujol, L.


    This article describes the development work on the porch RadSentry Canberra brand that uses 2 plastic scintillation detectors PVT large (172 x 32 x 4 cm). In this paper we present the different calibration necessary for the proper functioning of equipment: channel-energy calibration, energy resolution, and efficiency-energy.

  11. Use of a robotic manipulator in the simulation of the automation of a calibration process of dosemeters; Uso de un manipulador robotico en la simulacion de la automatizacion de un proceso de calibracion de dosimetros

    Benitez R, J.S.; Najera H, M.C. [Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, 11801 Mexico D.F. (Mexico)


    The development of a system based in a manipulative robot which simulates the operative sequence in a calibration process of dosemeters is presented. In this process it is performed the monitoring of the dosemeter positions and the calibrator by mean of an arm of articulated robot which develops the movement sequences and the taking a decision based on the information coming from the external sensors. (Author)

  12. Influence of geometry calibration in the determination of gamma emitters in active carbon filters; Influencia de la geometria de calibracion en la determinacion de emisores gamma en filtros de carbon activo

    Laurado Tarrago, M.; Tent Petrus, J.


    Testing laboratories participate in external quality controls. For these controls are useful it is necessary to eliminate factors that can introduce sources of variability in the results. A factor to consider in this paper as a source of variability is the calibration geometry chosen for the correct quantification of the activity concentrations. This work contains the results of the intercomparison exercise and from them analysed the effect of the geometry of calibration used in the quantification of activity concentrations present in the activated carbon filter. (Author)

  13. Methodology to include a correction for offset in the calibration of a Diode-based 2D verification device; Metodologia para incluir una correccion por offset en la calibracion de un dispositivo de verificacion 2D basado en diodos

    Ramirez Ros, J. C.; Pamos Urena, M.; Jerez Sainz, M.; Lobato Munoz, M.; Jodar Lopez, C. A.; Ruiz Lopez, M. a.; Carrasco Rodriguez, J. L.


    We propose a methodology to correct doses of device verification 2D MapChek2 planes by offset. This methodology provides an array of correction by Offset applied to the calibration per dose due to the Offset of the diode Central as well as the correction of the Offset of each diode on each acquisition. (Author)

  14. Calibration of a {sup 19} {sup 2}Ir source for high dose brachytherapy using various techniques; Calibracion de una fuente de {sup 192} Ir para braquiterapia de alta tasa de dosis mediante diversas tecnicas

    Montilla Prieto, Tedicel C., E-mail: tcdicel@gmaiLcam [Instituto de Oncologia Dr. Miguel Perez Carreno, Barbula (Venezuela, Bolivarian Republic of). Departamento de Fisica y Dosimetria; Padron Rivero, Alvaro D., E-mail: [Universidad de Carabobo, Barbula (Venezuela, Bolivarian Republic of). Facultad de Ciencias de la Salud. Departamento de Ciencias Fisiologicas


    In this research we studied three experimental procedures for calibration of a source of {sup 192}Ir to high dose rate for clinical brachytherapy use, and thus were compared and analysis of the advantages and disadvantages of each. For this study we quantified the value of the current kerma rate reference in air by three procedures: source calibration using a well chamber, with an cylindrical ionization chamber in air, and a cylindrical ionization chamber on a phantom, and this magnitude was compared with the value provided by the manufacturer of the source and thereby obtaining the deviation corresponding . Thus, it was found that the deviation corresponding to the source calibration making use of a well chamber, remained within tolerance, while the cylindrical ionization chamber in air and on phantom exceeded the standards established in some documents. However, although both the measurement in air and in the phantom are the procedures for the final calibration source, these can be used to verify that the delivered dose are in tolerance.

  15. Recognition of the dosimetric calibration capacities of Cuba by the International Bureau of Weights and Measures; Reconocimiento de las capacidades de calibracion dosimetrica de Cuba por el Buro Internacional de Pesas y Medidas

    Walwyn S, G.; Gutierrez L, S.; Tamayo G, J.A.; Gonzalez R, N.; Alonso V, G. [CPHR, Calle 20 No. 4113 e/ 41 y 47, Playa C.P. 11300, La Habana (Cuba)]. e-mail:


    The declared mission of the International Bureau of Weights and Measures are the world uniformity of the measurement, however until some years ago a formal mechanism didn't exist for its complete implementation. With this end arose the Mutual Recognition Agreement whose specific objective is to establish the grade of equivalence of the national standards, the one of mutually recognizing the calibration and measurement certificates and the one of providing to the governments of a sure technical tool in its commercial negotiations and regulatory matters at international level. Cuba like an associated country to the Meter Convention, signed the agreement and it intended to demonstrate the international equivalence of its standards. The best measurement and calibration capacities of the country in the dosimetric magnitudes are in the Secondary Laboratory of Dosimetric Calibration of the Protection and Hygiene of Radiations Center. This capacities were included in the Regional Metrological Organization COOMET in the year 2003. In June of the 2005 the metrological capacities have been approved and published in the databases of the International Bureau of Weights and Measures as demonstration of the high competition of the calibration works that its are carried out in the laboratory. This approval is one of the maximum international recognitions that the patterns of a country can receive and its are the result of 10 years of work of the laboratory like part of the international net OIEA/OMS, which has given it the possibility to gauge the patterns and of adopting internationally validated calibration methodologies. On the other hand, it has been decisive the participation of the laboratory in multiple international comparisons of their patterns, as well as the implementation of a system of administration of the quality credited by the competent national organ. The article reviews the technical work of the laboratory during several years that it gave as result this

  16. Secondary calibration laboratory for dosimetry in levels of therapy at the University of Santiago; Laboratorio secundario de calibracion para dosimetria en niveles de terapia en la Universidad de Santiago

    Gomez Rodriguez, F.; Gonzalez Castano, D. M.; Pazos Alvarez, A.


    A basic inherent benefits provided by the existence of a traceability chain radiation in any application, add the legal requirement for hospitals as pointed to by the RO. 1566/1998, which sets quality standards in radiotherapy. The decree attributed to hospital specialists radio physics in article 10 the responsibility for determining the acceptance and initial reference state of radiation generating equipment for therapeutic purposes, and the establishment and implementation of quality control programs associated and technical and physical aspects of radiation dosimetry. Different international organizations such as ICRU and IAEA recommendations on maintaining the accuracy of the dose delivered to patients in general, should be placed at least 5% considering the whole chain irradiation. In order to achieve this purpose it is necessary to establish programs of quality control and calibration dosimetric regular basis. The protocol of the IAEA TRS398 recommended dose calibration in water because it is a quantity of interest closest to clinical use and allows a relative uncertainty in the calibration environment reduced to 1%.. (Author)

  17. Use of calibration methodology of gamma cameras for the workers surveillance using a thyroid simulator; Uso de una metodologia de calibracion de camaras gamma para la vigilancia de trabajadores usando un simulador de tiroides

    Alfaro, M.; Molina, G.; Vazquez, R.; Garcia, O., E-mail: mercedes.alfaro@inin.gob.m [ININ, Carretera Mexico-Toluca s/n, 52750 Ocoyoacac, Estado de Mexico (Mexico)


    In Mexico there are a significant number of nuclear medicine centers in operation. For what the accidents risk related to the transport and manipulation of open sources used in nuclear medicine can exist. The National Institute of Nuclear Research (ININ) has as objective to establish a simple and feasible methodology for the workers surveillance related with the field of the nuclear medicine. This radiological surveillance can also be applied to the public in the event of a radiological accident. To achieve this it intends to use the available equipment s in the nuclear medicine centers, together with the neck-thyroid simulators elaborated by the ININ to calibrate the gamma cameras. The gamma cameras have among their component elements that conform spectrometric systems like the employees in the evaluation of the internal incorporation for direct measurements, reason why, besides their use for diagnostic for image, they can be calibrated with anthropomorphic simulators and also with punctual sources for the quantification of the radionuclides activity distributed homogeneously in the human body, or located in specific organs. Inside the project IAEA-ARCAL-RLA/9/049-LXXVIII -Procedures harmonization of internal dosimetry- where 9 countries intervened (Argentina, Brazil, Colombia, Cuba, Chile, Mexico, Peru, Uruguay and Spain). It was developed a protocol of cameras gamma calibration for the determination in vivo of radionuclides. The protocol is the base to establish and integrated network in Latin America to attend in response to emergencies, using nuclear medicine centers of public hospitals of the region. The objective is to achieve the appropriate radiological protection of the workers, essential for the sure and acceptable radiation use, the radioactive materials and the nuclear energy. (Author)

  18. System for calibration of instruments of x-ray measurement (CIR-X) applying the PGCS; Sistema para calibracion de instrumentos de medicion de rayos X (CIR-X) aplicando el PGCS

    Gaytan G, E.; Rivero G, T.; Cruz E, P.; Tovar M, V.M.; Vergara M, F.J. [ININ, 52750 La Marquesa, Estado de Mexico (Mexico)]. e-mail:


    The Department of Metrology of Ionizing Radiations of the ININ carries out calibration of instruments for X-ray measurement that determine the operation parameters in X-ray diagnostic machines of the health and private sectors. To facilitate this task, the Department of Automation and Instrumentation developed a system for acquisition and signals processing coming from a reference voltage divider with traceability at NIST that is connected directly to the X-rays tube. The system is integrated by the X-ray unit, the X-ray measurement equipment Dynalizer IIIU of RADCAL, a data acquisition card, a personal computer and the acquisition software and signals processing. (Author)

  19. Proposal of a calibration protocol of gamma chambers for estimation of the radionuclides incorporation in emergency situations; Propuesta de un protocolo de calibracion de camaras gamma para estimacion de la incorporacion de radionucleidos en situaciones de emergencia

    Dantas, B.M.; Lucena, E.; Dantas, A.L.A.; Araujo, F.; Melo, D. [Instituto de Radioprotecao e Dosimetria, Rio de Janeiro Brasil (Brazil); Teran, M.; Paolino, A. [Facultad de Quimica, Montevideo (Uruguay); Hermida, J.C. [Hospital de Clinicas, Facultad de Medicina, Montevideo (Uruguay); Rojo, A. [Autoridad Regulatoria Nuclear, Buenos Aires (Argentina); Puerta, J.A.; Morales, J. [Universidad Nacional de Colombia, Medellin (Colombia); Lopez B, G.M. [Centro de Proteccion e Higiene de las Radiaciones, Ciudad de la Habana (Cuba); Alfaro, M.; Ruiz, M.A. [Instituto Nacional de Investigaciones Nucleares (Mexico); Videla, R.; Pinones, O. [Comision Chilena de Energia Nuclear, Santiago (Chile); Gonzalez, S. [Instituto Peruano de Energia Nuclear, Lima (Peru); Navarro, T. [Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas, Madrid (Spain); Cruz S, R. [International Atomic Energy Agency, Vienna (Australia)


    In the last years in several countries has come increasing the concern with the possibility of accidents related to the transport and manipulation of open sources used in nuclear medicine. This carried out to the search of alternative methods for the monitoring of workers and individuals of the public exposed to the radionuclides incorporation like {sup 131} I, {sup 201} Tl, {sup 153} Sm among others. One of the options to assist the demand for monitoring of the radionuclides incorporation is the use of gamma chambers that are medical diagnostic equipment available in the own centers of nuclear medicine. The gamma chambers are used to obtain images of patient to which are administered a radionuclide of well-known activity with diagnostic purposes. These equipment have among its components elements that spectrometric systems like those used in the evaluation of the internal incorporation for direct measurements, reason why besides its use for diagnosis by image they can be gauged with anthropomorphic simulators and also with punctual sources for the quantification of the radionuclides activity, distributed homogeneously in the human body or located in specific organs. This work presents the development of a calibration protocol of gamma chambers for the in vivo determination of radionuclides and it proposes the implementation of the protocol in centers of nuclear medicine of the 9 countries participants of the project OAS-ARCAL-RLA/9/049-LXXVIII - Harmonization of procedures of internal dosimetry (Argentina, Brazil, Colombia, Cuba, Chile, Mexico, Peru, Uruguay and Spain). The protocol is the base to establish an integrated net to attend in the response to emergencies using nuclear medicine centers of public hospitals of the region. The proposal is an additional alternative for the monitoring of people in emergency situations where it is possible and feasible the use of the gamma chambers. This would avoid the person's transport with incorporation suspicion for

  20. Calibration of the High Resolution Gamma Spectrometry Detection Systems from the Environmental Radioactivity Unit and Radiological Surveillance (ARU and RV); Calibracion de los Sistemas Detectores de Espectrometria Gamma de Alta Resolucion de la Unidad de Radiactividad Ambiental y Vigilancia Radiologica (URA y VR)

    Suarez-Navarro, J. A.; Gasco, C.; Arana, M.; Suanez, A.


    In accordance with international regulations and with the quality criteria imposed by the internal quality system of the laboratory of gamma spectrometry, which belongs to the Unit of Environmental Radioactivity and Radiation Monitoring (URA y VR), it is necessary to perform a periodic calibration of the measurement systems of gamma spectrometry. The equipment calibration carries with the need for the preparation of the geometries of measurement and calibration in energy and in efficiencies. This report reflects the procedures for the preparation of the wide range of geometries available in the laboratory, as well as the different calibrations performed with those geometries in almost all the laboratory gamma detectors using the CANBERRA software Genie 2k. (Author) 9 refs.

  1. Calculation of dose absorbed for the verification of the calibration curves UH-Density electronic relative obtained with various mannequins; Calculo de dosis absorbida para la verificacion de las curvas de calibracion UH-Densidad electronica relativa obtenidas con distintos maniquies

    Perez Alvarez, M. E.; Sena Espinel, E. de; Delgado Aparicio, J. M.; Martin Rincon, C.; Garcia Repiso, S.; Ramos Pacho, J. A.; Verde Velasco, J. M.; Gomez Gonzalez, N.; Cons Perez, N.; Saez Beltran, M.


    In order to discern with what calibration curve is obtained a more accurate dosimetry distribution, dose measurements are carried out on the treatment unit and the values found are evaluated. (Author)

  2. Calibration of 232U solution; Calibracion de una disolucion de 232U

    Galan Valera, M. P.; Acena Berrenechea, M. L.


    A method for as certain the activity by alpha spectroscopy with semiconductor detectors, of a solution of 232U is presented. It consists of the comparison with a 233U solution activity previously measured in a gridded ionization chamber of 2{pi} geometry. The total measurement uncertainty is about + - 0,02. (Author) 9 refs.

  3. Flow measurement in a 170-MW hydraulic turbine using the Gibson method; Medicion del flujo de una turbina hidraulica de 170 MW utilizando el metodo Gibson

    Urquiza, Gustavo [Universidad Autonoma del Estado de Morelos (Mexico); Adamkowski, Adam [The Szewalski Institute of Fluid-Flow Machinery (Poland); Kubiak, Janusz; Sierra, Fernando [Universidad Autonoma del Estado de Morelos (Mexico); Janicki, Waldemar [The Szewalski Institute of Fluid-Flow Machinery (Poland); Fernandez, J. Manuel [Comision Federal de Electricidad (Mexico)


    This paper describes the methodology applied for measuring water flow through a 170-MW hydraulic turbine. The flow rate was measured using the pressure-time method, also known as the Gibson method. This method uses the well-known water hammer phenomenon in pipelines; in turbine penstocks, for instance. The version of this method used here is based on measuring, during total stop of the water stream, the time-history of pressure change in one section of the turbine penstock and relate it to the pressure in the upper reservoir to which the penstock is connected. The volumetric flow rate is determined from the relevant integration of the measured temporary pressure rise. Flow measurement was possible this way because the influence of the penstock inlet was negligible as far as an error of the measurement is concerned. The length of the penstock was 300 m. Previous experience and a standard IEC-41-1991 were the criteria adopted and applied. A fast and efficient acquisition system, including a 16 bit card, was used. The flow rate was calculated using a computer program developed and tested on several cases. The results obtained with the Gibson method were used for calibration of the on-line flow measuring system based on the Winter-Kennedy method as one of the index methods. This method is very often used for continuous monitoring of the flow rate through hydraulic turbines, when the calibration has been done on site by using the results of measurements obtained by the absolute method. Having measured the flow rate and output power, the efficiency was calculated for any operating conditions. A curve showing the best operating conditions based on the highest efficiency is presented and discussed. The details of the instrumentation, its installation, and the results obtained are discussed in the paper. [Spanish] Este articulo describe la metodologia aplicada para la medicion del flujo en una turbina hidraulica de 170 MW. El flujo se midio utilizando el metodo de presion

  4. Calibration and study of the measurement capabilities of real-time gamma spectrometry equipment developed for the renewal of the network of stations automatic monitoring environmental radiation of the Generalitat of Catalunya; Calibracion y estudio de las capacidades de medida de los equipos de espectrometria gamma en tiempo real desarrollados para la renovacion de la Red de Estaciones Automaticas de vigilancia radiologia ambiental de la Generalitat de Catalunya

    Casanovas, R.; Prieto, E.; Salvado, M.


    The renewal of the automatic environmental radiation surveillance network of the Generalitat of Catalunya has been carried out through the development and implementation of gamma-spectrometry-based equipment. the monitors use scintillation crystals, either Na(TI) or LaBr{sub 3}(Ce) and currently, there are 3 types of equipment water radioactivity monitors, aerosols on a particulate filter monitors and direct measurement monitors. In this paper, we expose the basic features of its operation, the details of their calibration and the minimum detectable activity concentration for some isotopes. (Author)

  5. Paleoclimatología y cronología del Würm reciente: un intento de síntesis



    Full Text Available RESUMEN: Las oscilaciones climáticas observadas en los núcleos polares (GRIP, GISP2 y la escala temporal que forman, pueden servirnos como marco de referencia general de las variaciones climáticas representadas en los yacimientos arqueológicos. De igual modo, los registros de núcleos marinos (MD95- 2042 y MD95-2043, junto con los terrestres del lago Grande de Monticchio y las dataciones U/Th de los espeleotemas de la cueva de Villars (SW de Francia, nos permiten comprobar la repercusión de estos cambios en nuestro continente. En su desarrollo, se hace evidente la necesidad de realizar calibraciones sobre las dataciones realizadas por el método 14C, lo que actualmente se está intentado por medio de síntesis sobre los datos recogidos en diversos núcleos sedimentarios, donde es posible conocer la datación de sus niveles por medios diferentes a los radiocarbónicos.ABSTRACT: Climatic oscillations determined in polar cores (GRIP, GISP2 and the chronological framework modeled in them can be used as sets of global references for determine climate oscillations represented in archaeological deposits. Record of marine cores (MD95-2042 and MD95-2043, with the terrestrial one of the Big Lake of Monticchio and the U/Th dates of Villars's cave stalactite (SW France, allow us to verify the impact of these changes in Europe. 14C dates need to be calibrated. In order to obtain a reliable calibration of them, nowadays it is in progress the build of framework synthesizing information gathered in those sedimentary cores containing levels that offer more dating possibilities that only radiocarbon methodology.

  6. Utilidad del dispositivo portátil PIKO6 para la detección precoz de EPOC en Atención Primaria

    Hidalgo Sierra, Vanesa


    [ES]Partiendo de la premisa que la EPOC es una enfermedad muy prevalente, pero infradiagnosticada, el presente trabajo ha consistido en la validación del dispositivo portátil Piko-6 para la detección precoz de EPOC en Atención Primaria, dado que el dispositivo portátil Piko-6 es mucho más barato, portátil y manejable, y no requiere adiestramiento especial por el personal sanitario ni de calibraciones periódicas. El objetivo principal de nuestro estudio fue validar la hipótesis de trabajo d...

  7. Influencia del momento de inercia del tambor y de diferentes ángulos de alimentación constantes sobre el calibre de las partículas de forraje procesado con picadores del tipo de tambor con alimentación manual. Parte I

    Pedro A. Valdés Hernández


    Full Text Available La investigación realizada, presenta como objetivo determinar la influencia del momento de inercia del tambor picador y de diferentes ángulos de alimentación constantes, sobre el calibre de las partículas de forraje procesado con máquinas picadoras de forraje de tallos gruesos del tipo de tambor con alimentación manual. En la investigación se aplica la modelación físico-matemática del proceso tecnológico, cuyos resultados se toman como base para el diseño de experimentos realizados durante el procesamiento de tallos de la variedad de Caña de Azúcar C323-68, siendo la de mayor uso en Cuba para la alimentación del ganado. La modelación físico matemática del proceso permite la predicción del calibre de las partículas con errores no superiores a 15,3%, con un nivel de significación del 1% para valores racionales del momento de inercia del tambor, calculado sobre bases teóricas. La racionalización del momento de inercia en la picadora de forraje de producción nacional MF-IIMA modelo EM-01 produjo un incremento en la calidad del desmenuzado, que se manifestó en un mayor porcentaje de partículas con calibres por debajo de 20 mm entre 43 y 48%. La modelación teórica resultó una valiosa herramienta de trabajo, posibilitando el perfeccionamiento del diseño de este tipo de picadora de forraje.

  8. Responsabilidad social: una ruta para alcanzar la excelencia en laboratorios de ensayos y calibraciones

    Herrera-Basurto, Raúl; Mercader-Trejo, F.; Torres Rivera, O.


    Se propone un modelo de gestión estratégico para incluir la gestión de la Responsabilidad Social (RS) como una herramienta que permita no sólo la responsabilidad por la calidad de los resultados y el compromiso con los clientes, la sociedad y los grupos de interés, sino también como una herramienta para lograr la excelencia de la organización. Una nueva concepción del laboratorio es una "condición absoluta" para su legitimidad hoy y mañana. El modelo de gestión es integrador y compatible, don...

  9. Calibration in water films GAFCHROMIC EBT radiochromic-2. Effects of Drying; Calibracion en agua de las peliculas radiocromicas Gafchromic EBT-2. Efectos de Secado

    Herrero, C.; Perez-Alija, J.; Alaman, C.; Almansa, J.; Vilches, M.


    Recent studies [1] show that immersion in water GAFCHROMIC EBT radiochromic film-2 has on these two effects: a progressive diffusion of water across its borders (fact which tells us the manufacturer and the effect of which we can avoid choosing a proper analysis region) and a uniform moisture through the polyester film, producing a variation in the measurement of optical density (OD) of the film. The latter effect is negligible for immersion times of less than 30 minutes. We study the effect of hydration and drying in these films when they have been submerged for 24 hours.

  10. Characterization and use of shadow cone in a calibration laboratory for neutron; Caracterizacion y empleo de conos de sombra en un laboratorio para calibracion neutronica

    Gallego Diaz, E.; Valle Diez, A. del; Lorente Fillol, A.; Vega Carrillo, H. R.


    We present the design, characterization and use of shadow cones neutron measurements of the Laboratory of Nuclear Engineering Department of the ETSII-UPM (LMN-UPM), with an irradiation room size 9 mx 16 mx 8 m in that used an {sup 2}41Am-Be source of 11 GBq (3 Ci).

  11. Calibración y Validación del Modelo SIMBA-POP en la Zona Bananera del Urabá Antioqueño, Colombia SIMBA-POP Model: Calibration and Validation in Urabá - Antioquia, Colombia

    Germán Wbeimar Guarín Giraldo


    Full Text Available Resumen. SIMBA-POP es un modelo desarrollado en Martinica (Francia, basado en los conceptos de cohortes y cadenas de cohortes de plantas, simula la dinámica poblacional de plantaciones de banano. El modelo original, modificado por Ochoa (2010, se programó en el software EVOLUCIÓN 4.0, desarrollado por el grupo SIMON de la Universidad Industrial de Santander y se calibró y validó para las condiciones de temperatura media habituales en el Urabá Antioqueño; la más importante zona productora de banano en Colombia. En el proceso de calibración se ajustan los ciclos de embolse y cosecha, así como los tiempos en que ocurren los máximos valores de embolse. Para controlar la calibración se usó la desviación media (DM, el error medio de predicción (EMP, el error cuadrático medio de predicción (ECMP, la raíz del error cuadrático medio (RECMP, curvas de masa y coeficiente de correlación lineal. Los resultados mostraron que SIMBA-POP es capaz de simular de forma verosímil la dinámica del ciclo anual de producción en Urabá; particularmente reproduce bien los tiempos de los picos y los valles de producción, aunque la magnitud de estos valles y picos es altamente sensible a los parámetros del modelo y condiciones climáticas no consideradas.Abstract. SIMBA - POP is a model developed in Martinique (France, which uses the concepts of cohorts and cohort chains to simulate the population dynamics of a commercial plantation of bananas. The original model was slightly modified by Ochoa (2010 in order to correct for an error in the flowering rate curve. The model was programmed in EVOLUCION 4.0. SIMBA - POP was then calibrated and validated for the normal climatic pattern in Urabá (Antioquia. Care was taken that the model adequately represent bagging and harvesting cycles, and the time at which maximum values of bagging (flowering were reached. To evaluate the calibration we used average deviation (DM, the average error of prediction (EMP

  12. Soil Moisture Measurement through Time Domain Reflectometry (TDR). Irrigation Application; Medicion de la humedad del suelo por reflectometria en el dominio temporal (TDR). Aplicacion en irrigacion

    Quinonez Pedroza, Hector E.; Ruelle, Pierre; Nemeth Ildiko [Cemagre, (France)


    Knowledge of how moisture varies in the soil is essential to determine the actual rate of evapotranspiration in crops and, thus, the amount of water necessary for irrigation. Lisimeters, tensiometers and the neutron probe have been widely used for precise measurement of soil moisture in order to calibrate models of irrigation scheduling or to calculate volumes of irrigation water. Despite the usefulness of such measurements, these methods have limited use because of problems like high cost of lisimeters and neutron probes as well as their strict regulation, limited range of validity of tensiometers. Moreover, the measurements obtained from these methods generally describe noncontinuous points in time. Time Domain Reflectometry (TDR) is a viable to such methods because of its precision and the continuity of its measurements. This study evaluates soil moisture, using the neutron and TDR methods in maize crop irrigated by gravity in closed furrows. The results show that even though the absolute determinations of moisture are different in the 30-90 cm profiles, moisture variations were similar in all cases and respond consequently to the additions of water and to evapotranspiration. Neutron probe drawback become evident on its no-continuous measurements contrasting with the continuous measurements of TDR, which allows a detailed analysis of the infiltration and evapotranspiration phenomenal at instantaneous time scales. [Spanish] El conocimiento de la variacion de la humedad del suelo es esencial para determinar la evapotranspiracion real de los cultivos. Esta ultima, a su vez, permite conocer las cantidades necesarias de agua de riego. Los lisimetros, los tensiometros y el aspersor de nuestro s han sido ampliamente usados para la medicion precisa de la humedad del suelo con propositos de calibracion de modelos de programacion del riego o para su utilizacion directa en el calculo de volumenes de riego. A pesar de la importancia de tales mediciones, su uso es muy

  13. Use of Near Infrared Reflectance (NIR Spectroscopy to Predict Chemical Composition of Forages in Broad-Based Calibration Models Uso de la Espectroscopía de Reflectancia en el Infrarrojo Cercano (NIR Para Predecir la Composición Química de Forrajes en Modelos de Calibración Amplia

    Jaime Garcia


    Full Text Available The objective of the study was to evaluate the potential of near infrared reflectance (NIR spectroscopy as a rapid method to predict the chemical composition of forage in broad-based calibration models. In total, 650 samples representing a wide range of chemical characteristics, phenological states and origins were scanned in an NIR instrument. The coefficient of determination in calibration (R² and standard error in cross validation (SECV for the NIR calibration models were as follows: dry matter 0.95 (SECV: 0.7%, crude protein 0.98 (SECV: 0.98%, ash 0.90 (SECV: 0.99%, in vitro organic matter digestibility 0.90 (SECV: 3.6%, acid detergent fiber 0.95 (SECV: 2.0% and neutral detergent fiber 0.86 (SECV: 5.4% on a dry matter basis. The results demonstrated the potential of NIR to predict the chemical composition of different forage plant species ; however, it is suggested that the technique could be used as a routine procedure to apply in breeding programs only if calibration is done for each species, season and particular conditions.El objetivo del trabajo fue evaluar el potencial de la espectroscopía de reflectancia en el infrarrojo cercano (NIR como método rápido para la predicción de la composición química de forrajes en modelos de calibración amplia. Un total de 650 muestras de forrajes de un amplio rango de características químicas, origen y estados fenólogicos, fueron analizadas en un instrumento NIR en reflectancia. Los coeficientes de determinación en calibración (R² y los errores estándares de la validación cruzada (SECV para las calibraciones NIR fueron 0,95 (SECV: 0,7%, 0,98 (SECV: 0,98%, 0,90 (SECV: 0,99%, 0,95 (SECV: 2,0%, 0,86 (SECV: 5,4% y 0,90 (SECV: 3,6% para materia seca (MS, proteína cruda (PC, cenizas, fibra detergente ácido (FDA, fibra detergente neutro (FDN, y digestibilidad in vitro de la materia orgánica (DIVMO, base materia seca, respectivamente. Los resultados demostraron el potencial del NIR para

  14. Calibration uncertainty

    Heydorn, Kaj; Anglov, Thomas


    Methods recommended by the International Standardization Organisation and Eurachem are not satisfactory for the correct estimation of calibration uncertainty. A novel approach is introduced and tested on actual calibration data for the determination of Pb by ICP-AES. The improved calibration...... uncertainty was verified from independent measurements of the same sample by demonstrating statistical control of analytical results and the absence of bias. The proposed method takes into account uncertainties of the measurement, as well as of the amount of calibrant. It is applicable to all types of...

  15. Camera calibration

    Andrade-Cetto, J.


    This report is a tutorial on pattern based camera calibration for computer vision. The methods presented here allow for the computation of the intrinsic and extrinsic parameters of a camera. These methods are widely available in the literature, and they are only summarized here as an easy and comprehensive reference for researchers at the Institute and their collaborators.

  16. Problema de calibración de mercado y estructura implícita del modelo de bonos de Black-Cox || Market Calibration Problem and the Implied Structure of the Black-Cox Bond Model

    Santana Jiménez, Lisette Josefina


    Full Text Available El principal resultado de este artículo consiste en la resolución del problema inverso del modelo de Black-Cox (1976, usando el método propuesto por Sukhomlin (2007. Se parte del enfoque retrógrado (backward para obtener una expresión exacta de la volatilidad implícita en función de parámetros cuantificables con datos de mercado y de variables conocidas. Se descubre la existencia de dos valores de la volatilidad para un activo subyacente en elmodelo referido, lo que indica que las asunciones tradicionales no lo definen de manera unívoca. Se encuentra la causa de que el modelo de Black-Cox contenga dos valores de la volatilidad. Además, se lleva a cabo una simulación, afin de verificar, numéricamente, que la expresión obtenida para la volatilidad es la inversión de la fórmula que representa la probabilidad de que la firma no alcance un nivel de insolvencia antes del tiempo de madurez de la deuda. Finalmente, se resuelve el problema de calibración de mercadodesde el punto de vista directo (forward, encontrándose una expresión que resulta de mayor utilidad para los agentes de mercado. || The main result of this paper consists in the resolution of the inverse problem for the Black-Cox (1976 model, using the method proposed by Sukhomlin (2007. Based on the backward approach, we obtain an exact expression of the implied volatility expressed as a function of quantifiable market parameters and known variables. We discover the existence of two values of the volatility for an underlying asset, in the referred model, which means that the model's traditional assumptions do not define it univocally. We findthe cause that the Black-Cox model contains two values of the volatility. Besides, we carry out a simulation in order to verify, numerically, that our volatility expression is in fact the inversion of the formula that represents the probability that the firm has not reached the reorganization boundary before the debt expires. Finally, we

  17. Calibration Binaries

    Drummond, J.


    Two Excel Spreadsheet files are offered to help calibrate telescope or camera image scale and orientation with binary stars for any time. One is a personally selected list of fixed position binaries and binaries with well-determined orbits, and the other contains all binaries with published orbits. Both are derived from the web site of the Washington Double Star Library. The spreadsheets give the position angle and separation of the binaries for any entered time by taking advantage of Excel's built in iteration function to solve Kepler's transcendental equation.

  18. ALTEA calibration

    Zaconte, V.; Altea Team

    The ALTEA project is aimed at studying the possible functional damages to the Central Nervous System (CNS) due to particle radiation in space environment. The project is an international and multi-disciplinary collaboration. The ALTEA facility is an helmet-shaped device that will study concurrently the passage of cosmic radiation through the brain, the functional status of the visual system and the electrophysiological dynamics of the cortical activity. The basic instrumentation is composed by six active particle telescopes, one ElectroEncephaloGraph (EEG), a visual stimulator and a pushbutton. The telescopes are able to detect the passage of each particle measuring its energy, trajectory and released energy into the brain and identifying nuclear species. The EEG and the Visual Stimulator are able to measure the functional status of the visual system, the cortical electrophysiological activity, and to look for a correlation between incident particles, brain activity and Light Flash perceptions. These basic instruments can be used separately or in any combination, permitting several different experiments. ALTEA is scheduled to fly in the International Space Station (ISS) in November, 15th 2004. In this paper the calibration of the Flight Model of the silicon telescopes (Silicon Detector Units - SDUs) will be shown. These measures have been taken at the GSI heavy ion accelerator in Darmstadt. First calibration has been taken out in November 2003 on the SDU-FM1 using C nuclei at different energies: 100, 150, 400 and 600 Mev/n. We performed a complete beam scan of the SDU-FM1 to check functionality and homogeneity of all strips of silicon detector planes, for each beam energy we collected data to achieve good statistics and finally we put two different thickness of Aluminium and Plexiglas in front of the detector in order to study fragmentations. This test has been carried out with a Test Equipment to simulate the Digital Acquisition Unit (DAU). We are scheduled to

  19. Trinocular Calibration Method Based on Binocular Calibration

    CAO Dan-Dan; Luo, Chun; GAO Shu-Yuan; Wang, Yun; Li, Wen-Bin; XU Zhen-Ying


    In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error) of the global calibration of ...

  20. Traceable Pyrgeometer Calibrations

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina


    This poster presents the development, implementation, and operation of the Broadband Outdoor Radiometer Calibrations (BORCAL) Longwave (LW) system at the Southern Great Plains Radiometric Calibration Facility for the calibration of pyrgeometers that provide traceability to the World Infrared Standard Group.

  1. Calibration curve to establish the exposure dose at Co{sup 60} gamma radiation; Curva de calibracion para establecer dosis de exposicion a radiacion gamma de Co{sup 60}

    Guerrero C, C.; Brena V, M. [Departamento de Biologia, Instituto Nacional de Investigaciones Nucleares, A.P. 18-1027, Mexico D.F. (Mexico)


    The biological dosimetry is an adequate method for the dose determination in cases of overexposure to ionizing radiation or doubt of the dose obtained by physical methods. It is based in the aberrations analysis produced in the chromosomes. The behavior of leisure in chromosomes is of dose-response type and it has been generated curves in distinct laboratories. Next is presented the curve for gamma radiation produced in the National Institute of Nuclear Research (ININ) laboratory. (Author)

  2. Radiological protection and calibration of an activity meter with cesium and barium sources in a nuclear medicine center; Proteccion radiologica y calibracion de un activimetro con fuentes de cesio y bario en un centro de medicina nuclear

    Morales L, M.E. [IPEN INEN Centro de Medicina Nuclear, Av. Aviacion 3799, Lima 34 (Peru)


    Presently work is shown the results when gauging a team Deluxe Isotope (Caliper 11) with some sources of Cesium 137 and Barium 133, in a Center of Medicine Nuclear that operates from the anus 1983 in a modern building inside the one Institute of Illnesses Neoplasia (Inn). This Center was equipped initially with teams donated by the International Organism of Atomic Energy (Oa) with those that it develops assistance, educational works and of investigation, giving services to patient of the Inn and other public and private medical centers. (Author)

  3. Trinocular Calibration Method Based on Binocular Calibration

    CAO Dan-Dan


    Full Text Available In order to solve the self-occlusion problem in plane-based multi-camera calibration system and expand the measurement range, a tri-camera vision system based on binocular calibration is proposed. The three cameras are grouped into two pairs, while the public camera is taken as the reference to build the global coordinate. By calibration of the measured absolute distance and the true absolute distance, global calibration is realized. The MRE (mean relative error of the global calibration of the two camera pairs in the experiments can be as low as 0.277% and 0.328% respectively. Experiment results show that this method is feasible, simple and effective, and has high precision.

  4. Automated Camera Calibration

    Chen, Siqi; Cheng, Yang; Willson, Reg


    Automated Camera Calibration (ACAL) is a computer program that automates the generation of calibration data for camera models used in machine vision systems. Machine vision camera models describe the mapping between points in three-dimensional (3D) space in front of the camera and the corresponding points in two-dimensional (2D) space in the camera s image. Calibrating a camera model requires a set of calibration data containing known 3D-to-2D point correspondences for the given camera system. Generating calibration data typically involves taking images of a calibration target where the 3D locations of the target s fiducial marks are known, and then measuring the 2D locations of the fiducial marks in the images. ACAL automates the analysis of calibration target images and greatly speeds the overall calibration process.

  5. ORNL calibrations facility

    The ORNL Calibrations Facility is operated by the Instrumentation Group of the Industrial Safety and Applied Health Physics Division. Its primary purpose is to maintain radiation calibration standards for calibration of ORNL health physics instruments and personnel dosimeters. This report includes a discussion of the radioactive sources and ancillary equipment in use and a step-by-step procedure for calibration of those survey instruments and personnel dosimeters in routine use at ORNL

  6. Analytical multicollimator camera calibration

    Tayman, W.P.


    Calibration with the U.S. Geological survey multicollimator determines the calibrated focal length, the point of symmetry, the radial distortion referred to the point of symmetry, and the asymmetric characteristiecs of the camera lens. For this project, two cameras were calibrated, a Zeiss RMK A 15/23 and a Wild RC 8. Four test exposures were made with each camera. Results are tabulated for each exposure and averaged for each set. Copies of the standard USGS calibration reports are included. ?? 1978.

  7. Spiral reader calibration

    The method to calibrate the spiral reader (SR) is presented. A brief description of the main procedures of the calibration program SCALP, adapted for the IHEP equipment and purposes, is described. The precision characteristics of the IHEP SR have been analysed on the results, presented in the form of diagrams. There is a calibration manual for the user

  8. Residual gas analyzer calibration

    Lilienkamp, R. H.


    A technique which employs known gas mixtures to calibrate the residual gas analyzer (RGA) is described. The mass spectra from the RGA are recorded for each gas mixture. This mass spectra data and the mixture composition data each form a matrix. From the two matrices the calibration matrix may be computed. The matrix mathematics requires the number of calibration gas mixtures be equal to or greater than the number of gases included in the calibration. This technique was evaluated using a mathematical model of an RGA to generate the mass spectra. This model included shot noise errors in the mass spectra. Errors in the gas concentrations were also included in the valuation. The effects of these errors was studied by varying their magnitudes and comparing the resulting calibrations. Several methods of evaluating an actual calibration are presented. The effects of the number of gases in then, the composition of the calibration mixture, and the number of mixtures used are discussed.

  9. Separation of 210Pb, 210Bi and 210Po by ion exchange and their Iiquid scintillation standardization; Separacion del 210Pb, 210Bi y 2I0Po mediante columna de cambio ionico y su calibracion por centelleo liquido

    Rodriguez, L.; Jimenez, A.; Grau, A.


    We applied the CIEMAT/NIST method and alpha/beta discrimination to ''210Pb samples in equilibrium with its daughters, by preparing homogeneous and gel samples. The stability of samples was tested in different available cocktails, HiSafe''TM II, HiSafe''TM III, Ultima-Gold''TM, Ultima-Gold''TM XR, Ultima-Gold''TM AB, Insta-Gel''R and e Insta-Gel''R lI. Also we analyzed the disequilibrium of the radioactive chain 210Pb+210Bi+210Po, achieving an excellent agreement between the results of the spectrum unfolding method and the experimental values. (Author) 13 refs.

  10. Preparation and LSC Standardization of ''89Sr (DNP) Using the CIEMAT/NIST Method; Preparacion del ''89Sr(DNP) y calibracion por centelleo liquido, mediante el metodo CIEMAT/NIST

    Rodriguez Barquero, L.; Los Arcos Merino, J. M.; Grau Malonda, A.


    A procedure for preparation of liquid scintillation counting samples of the strontium DNP complex, labelled with ''89Sr, is described, the chemical quench, the counting stability and spectral evolution of this compound is studied in six scintillators, Toluene, Toluene-alcohol, Dioxane-naphthalene, HiSafe II, Ultima- Gold and Instagel. The liquid scintillation standardization of 89Sr-DNP by the CIEMAT/NIST method, using HiSafe II and Ultima-Gold scintillators, has been carried out. The discrepancies between experimental and computed efficiencies are lower than 0.38% and 0.48%, respectively. The solution has been standardized in terms of activity concentration to an overall uncertainty of 0,38%. (Author) 10 refs.

  11. RF impedance measurement calibration

    The intent of this note is not to explain all of the available calibration methods in detail. Instead, we will focus on the calibration methods of interest for RF impedance coupling measurements and attempt to explain: (1). The standards and measurements necessary for the various calibration techniques. (2). The advantages and disadvantages of each technique. (3). The mathematical manipulations that need to be applied to the measured standards and devices. (4). An outline of the steps needed for writing a calibration routine that operated from a remote computer. For further details of the various techniques presented in this note, the reader should consult the references

  12. The COS Calibration Pipeline

    Hodge, Philip E.; Keyes, C.; Kaiser, M.


    The COS calibration pipeline (CALCOS) includes three main components: basic calibration, wavelength calibration, and spectral extraction. Calibration of modes using the far ultraviolet (FUV) and near ultraviolet (NUV) detectors share a common structure, although the individual reference files differ and there are some additional steps for the FUV channel. The pipeline is designed to calibrate data acquired in either ACCUM or time-tag mode. The basic calibration includes pulse-height filtering and geometric correction for FUV, and flat-field, deadtime, and Doppler correction for both detectors. Wavelength calibration can be done either by using separate lamp exposures or by taking several short lamp exposures concurrently with a science exposure. For time-tag data, the latter mode ("tagflash") will allow better correction of potential drift of the spectrum on the detector. One-dimensional spectra will be extracted and saved in a FITS binary table. Separate columns will be used for the flux-calibrated spectrum, error estimate, and the associated wavelengths. CALCOS is written in Python, with some functions in C. It is similar in style to other HST pipeline code in that it uses an association table to specify which files to be included, and the calibration steps to be performed and the reference files to use are specified by header keywords. Currently, in conjunction with the Instrument Definition Team (led by J. Green), the ground-based reference files are being refined, delivered, and tested with the pipeline.

  13. Energy calibration via correlation

    Maier, Daniel; Limousin, Olivier


    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be less than ~ 0.1 keV. Energy calibration via correlation can be applied to any kind of calibration spectra and shows a robust behavior at low counting statistics. It enables a fast and accurate calibration that can be used to monitor the spectroscopic properties of a detector system in near realtime.

  14. Photogrammetric camera calibration

    Tayman, W.P.; Ziemann, H.


    Section 2 (Calibration) of the document "Recommended Procedures for Calibrating Photogrammetric Cameras and Related Optical Tests" from the International Archives of Photogrammetry, Vol. XIII, Part 4, is reviewed in the light of recent practical work, and suggestions for changes are made. These suggestions are intended as a basis for a further discussion. ?? 1984.

  15. Sandia WIPP calibration traceability

    Schuhen, M.D. [Sandia National Labs., Albuquerque, NM (United States); Dean, T.A. [RE/SPEC, Inc., Albuquerque, NM (United States)


    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities.

  16. The GERDA calibration system

    A system with three identical custom made units is used for the energy calibration of the GERDA Ge diodes. To perform a calibration the 228Th sources are lowered from the parking positions at the top of the cryostat. Their positions are measured by two independent modules. One, the incremental encoder, counts the holes in the perforated steel band holding the sources, the other measures the drive shaft's angular position even if not powered. The system can be controlled remotely by a Labview program. The calibration data is analyzed by an iterative calibration algorithm determining the calibration functions for different energy reconstruction algorithms and the resolution of several peaks in the 228Th spectrum is determined. A Monte Carlo simulation using the GERDA simulation software MAGE has been performed to determine the background induced by the sources in the parking positions.

  17. Sandia WIPP calibration traceability

    This report summarizes the work performed to establish calibration traceability for the instrumentation used by Sandia National Laboratories at the Waste Isolation Pilot Plant (WIPP) during testing from 1980-1985. Identifying the calibration traceability is an important part of establishing a pedigree for the data and is part of the qualification of existing data. In general, the requirement states that the calibration of Measuring and Test equipment must have a valid relationship to nationally recognized standards or the basis for the calibration must be documented. Sandia recognized that just establishing calibration traceability would not necessarily mean that all QA requirements were met during the certification of test instrumentation. To address this concern, the assessment was expanded to include various activities

  18. Remote calibration of ionization chambers for radioactivity calibration

    A new calibration technique, referred to as e-trace, has been developed by the National Institute of Advanced Industrial Science and Technology (AIST). The e-trace technique enables rapid remote calibration of measurement equipment and requires minimal resources. We calibrated radioisotope calibrators of the Japan Radioisotope Association (JRIA) and the Nishina Memorial Cyclotron Center (NMCC) remotely and confirmed that remote calibration provided results that are consistent with the results obtained by existing methods within the limits of uncertainty. Accordingly, e-trace has been approved as the standard calibration method at AIST. We intend to apply remote calibration to radioisotope calibrators in hospitals and isotope facilities. (author)

  19. Calibrations of photomultiplier tubes

    The experimental methods for calibration photomultiplier tubes used in the multichannel fast-pulse-detection system of Thomson scattering measurements for nuclear fusion devices is reported. The most important parameters of the photomultiplier tubes to be calibrated include: linearity of output electric signals to input light signals, response time of pulsed light, spectral response, absolute responsibility, and sensitivity as a function of the chain voltage. The calibrations of all these parameters are carried out by using EMI 9558 B and RCA 7265 photomultiplier tubes respectively. The experimental methods presented in the paper are common to those quantitative measurements that require phomultiplier tubes as detectors

  20. Equipment for dosemeter calibration

    The device is used for precise calibration of dosimetric instrumentation, such as used at nuclear facilities. The high precision of the calibration procedure is primarily due to the fact that one single and steady radiation source is used. The accurate alignment of the source and the absence of shielding materials in the beam axis make for high homogeneity of the beam and reproducibility of the measurement; this is also contributed to by the horizontal displacement of the optical bench, which ensures a constant temperature field and the possibility of adjusting the radiation source at a sufficient distance from the instrument to be calibrated. (Z.S.). 3 figs

  1. Lidar Calibration Centre

    Pappalardo, Gelsomina; Freudenthaler, Volker; Nicolae, Doina; Mona, Lucia; Belegante, Livio; D'Amico, Giuseppe


    This paper presents the newly established Lidar Calibration Centre, a distributed infrastructure in Europe, whose goal is to offer services for complete characterization and calibration of lidars and ceilometers. Mobile reference lidars, laboratories for testing and characterization of optics and electronics, facilities for inspection and debugging of instruments, as well as for training in good practices are open to users from the scientific community, operational services and private sector. The Lidar Calibration Centre offers support for trans-national access through the EC HORIZON2020 project ACTRIS-2.

  2. SPOTS Calibration Example

    Patterson E.


    Full Text Available The results are presented using the procedure outlined by the Standardisation Project for Optical Techniques of Strain measurement to calibrate a digital image correlation system. The process involves comparing the experimental data obtained with the optical measurement system to the theoretical values for a specially designed specimen. The standard states the criteria which must be met in order to achieve successful calibration, in addition to quantifying the measurement uncertainty in the system. The system was evaluated at three different displacement load levels, generating strain ranges from 289 µstrain to 2110 µstrain. At the 289 µstrain range, the calibration uncertainty was found to be 14.1 µstrain, and at the 2110 µstrain range it was found to be 28.9 µstrain. This calibration procedure was performed without painting a speckle pattern on the surface of the metal. Instead, the specimen surface was prepared using different grades of grit paper to produce the desired texture.

  3. Traceable Pyrgeometer Calibrations

    Dooraghi, Mike; Kutchenreiter, Mark; Reda, Ibrahim; Habte, Aron; Sengupta, Manajit; Andreas, Afshin; Newman, Martina; Webb, Craig


    This presentation provides a high-level overview of the progress on the Broadband Outdoor Radiometer Calibrations for all shortwave and longwave radiometers that are deployed by the Atmospheric Radiation Measurement program.

  4. Air Data Calibration Facility

    Federal Laboratory Consortium — This facility is for low altitude subsonic altimeter system calibrations of air vehicles. Mission is a direct support of the AFFTC mission. Postflight data merge is...

  5. Approximation Behooves Calibration

    da Silva Ribeiro, André Manuel; Poulsen, Rolf


    Calibration based on an expansion approximation for option prices in the Heston stochastic volatility model gives stable, accurate, and fast results for S&P500-index option data over the period 2005–2009.......Calibration based on an expansion approximation for option prices in the Heston stochastic volatility model gives stable, accurate, and fast results for S&P500-index option data over the period 2005–2009....

  6. Scanner calibration revisited

    Pozhitkov Alexander E


    Full Text Available Abstract Background Calibration of a microarray scanner is critical for accurate interpretation of microarray results. Shi et al. (BMC Bioinformatics, 2005, 6, Art. No. S11 Suppl. 2. reported usage of a Full Moon BioSystems slide for calibration. Inspired by the Shi et al. work, we have calibrated microarray scanners in our previous research. We were puzzled however, that most of the signal intensities from a biological sample fell below the sensitivity threshold level determined by the calibration slide. This conundrum led us to re-investigate the quality of calibration provided by the Full Moon BioSystems slide as well as the accuracy of the analysis performed by Shi et al. Methods Signal intensities were recorded on three different microarray scanners at various photomultiplier gain levels using the same calibration slide from Full Moon BioSystems. Data analysis was conducted on raw signal intensities without normalization or transformation of any kind. Weighted least-squares method was used to fit the data. Results We found that initial analysis performed by Shi et al. did not take into account autofluorescence of the Full Moon BioSystems slide, which led to a grossly distorted microarray scanner response. Our analysis revealed that a power-law function, which is explicitly accounting for the slide autofluorescence, perfectly described a relationship between signal intensities and fluorophore quantities. Conclusions Microarray scanners respond in a much less distorted fashion than was reported by Shi et al. Full Moon BioSystems calibration slides are inadequate for performing calibration. We recommend against using these slides.

  7. Calibrating nacelle lidars

    Courtney, M.


    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report presents four concepts for performing such a nacelle lidar calibration. Of the four methods, two are found to be immediately relevant and are pursued in some detail. The first of these is a line of sight calibration method in which both lines of sight (for a two beam lidar) are individually calibrated by accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a representative distribution of radial wind speeds. An alternative method is to place the nacelle lidar on the ground and incline the beams upwards to bisect a mast equipped with reference instrumentation at a known height and range. This method will be easier and faster to implement and execute but the beam inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work. (Author)

  8. Energy calibration via correlation

    Maier, Daniel


    The main task of an energy calibration is to find a relation between pulse-height values and the corresponding energies. Doing this for each pulse-height channel individually requires an elaborated input spectrum with an excellent counting statistics and a sophisticated data analysis. This work presents an easy to handle energy calibration process which can operate reliably on calibration measurements with low counting statistics. The method uses a parameter based model for the energy calibration and concludes on the optimal parameters of the model by finding the best correlation between the measured pulse-height spectrum and multiple synthetic pulse-height spectra which are constructed with different sets of calibration parameters. A CdTe-based semiconductor detector and the line emissions of an 241 Am source were used to test the performance of the correlation method in terms of systematic calibration errors for different counting statistics. Up to energies of 60 keV systematic errors were measured to be le...

  9. Calibration of sound velocimeter in pure water

    Li, Zhiwei; Zhang, Baofeng; Li, Tao; Zhu, Junchao; Xie, Ziming


    Accurate measurement of sound speed is important to calibrate a sound velocity profiler which provides real-time sound velocity to the sonar equipment in oceanographic survey. The sound velocity profiler calculates the sound speed by measuring the time-of-flight of a 1 MHz single acoustic pulse to travel over about 300 mm path. A standard sound velocimeter instrument was invited to calibrate the sound velocity profiler in pure water at temperatures of 278,283, 288, 293, 298, 303 and 308K in a thermostatic vessel at one atmosphere. The sound velocity profiler was deployed in the thermostatic vessel alongside the standard sound velocimeter instrument and two platinum resistance thermometers (PRT) which were calibrated to 0.002k by comparison with a standard PRT. Time of flight circuit board was used to measure the time-of-flight to 22 picosecond precision. The sound speed which was measured by the sound velocity profiler was compared to the standard sound speed calculated by UNESCO to give the laboratory calibration coefficients and was demonstrated agreement with CTD-derived sound speed using Del Grosso's seawater equation after removing a bias.

  10. HAWC Timing Calibration

    Huentemeyer, Petra; Dingus, Brenda


    The High-Altitude Water Cherenkov (HAWC) Experiment is a second-generation highsensitivity gamma-ray and cosmic-ray detector that builds on the experience and technology of the Milagro observatory. Like Milagro, HAWC utilizes the water Cherenkov technique to measure extensive air showers. Instead of a pond filled with water (as in Milagro) an array of closely packed water tanks is used. The event direction will be reconstructed using the times when the PMTs in each tank are triggered. Therefore, the timing calibration will be crucial for reaching an angular resolution as low as 0.25 degrees.We propose to use a laser calibration system, patterned after the calibration system in Milagro. Like Milagro, the HAWC optical calibration system will use ~1 ns laser light pulses. Unlike Milagro, the PMTs are optically isolated and require their own optical fiber calibration. For HAWC the laser light pulses will be directed through a series of optical fan-outs and fibers to illuminate the PMTs in approximately one half o...

  11. Calibration Under Uncertainty.

    Swiler, Laura Painton; Trucano, Timothy Guy


    This report is a white paper summarizing the literature and different approaches to the problem of calibrating computer model parameters in the face of model uncertainty. Model calibration is often formulated as finding the parameters that minimize the squared difference between the model-computed data (the predicted data) and the actual experimental data. This approach does not allow for explicit treatment of uncertainty or error in the model itself: the model is considered the %22true%22 deterministic representation of reality. While this approach does have utility, it is far from an accurate mathematical treatment of the true model calibration problem in which both the computed data and experimental data have error bars. This year, we examined methods to perform calibration accounting for the error in both the computer model and the data, as well as improving our understanding of its meaning for model predictability. We call this approach Calibration under Uncertainty (CUU). This talk presents our current thinking on CUU. We outline some current approaches in the literature, and discuss the Bayesian approach to CUU in detail.

  12. GTC Photometric Calibration

    di Cesare, M. A.; Hammersley, P. L.; Rodriguez Espinosa, J. M.


    We are currently developing the calibration programme for GTC using techniques similar to the ones use for the space telescope calibration (Hammersley et al. 1998, A&AS, 128, 207; Cohen et al. 1999, AJ, 117, 1864). We are planning to produce a catalogue with calibration stars which are suitable for a 10-m telescope. These sources will be not variable, non binary and do not have infrared excesses if they are to be used in the infrared. The GTC science instruments require photometric calibration between 0.35 and 2.5 microns. The instruments are: OSIRIS (Optical System for Imaging low Resolution Integrated Spectroscopy), ELMER and EMIR (Espectrógrafo Multiobjeto Infrarrojo) and the Acquisition and Guiding boxes (Di Césare, Hammersley, & Rodriguez Espinosa 2005, RevMexAA Ser. Conf., 24, 231). The catalogue will consist of 30 star fields distributed in all of North Hemisphere. We will use fields containing sources over the range 12 to 22 magnitude, and spanning a wide range of spectral types (A to M) for the visible and near infrared. In the poster we will show the method used for selecting these fields and we will present the analysis of the data on the first calibration fields observed.

  13. Calibrating nacelle lidars

    Courtney, Michael

    Nacelle mounted, forward looking wind lidars are beginning to be used to provide reference wind speed measurements for the power performance testing of wind turbines. In such applications, a formal calibration procedure with a corresponding uncertainty assessment will be necessary. This report...... accurately aligning the beam to pass close to a reference wind speed sensor. A testing procedure is presented, reporting requirements outlined and the uncertainty of the method analysed. It is seen that the main limitation of the line of sight calibration method is the time required to obtain a...... inclination introduces extra uncertainties. A procedure for conducting such a calibration is presented and initial indications of the uncertainties given. A discussion of the merits and weaknesses of the two methods is given together with some proposals for the next important steps to be taken in this work....


    L. Barazzetti


    Full Text Available In photogrammetry a camera is considered calibrated if its interior orientation parameters are known. These encompass the principal distance, the principal point position and some Additional Parameters used to model possible systematic errors. The current state of the art for automated camera calibration relies on the use of coded targets to accurately determine the image correspondences. This paper presents a new methodology for the efficient and rigorous photogrammetric calibration of digital cameras which does not require any longer the use of targets. A set of images depicting a scene with a good texture are sufficient for the extraction of natural corresponding image points. These are automatically matched with feature-based approaches and robust estimation techniques. The successive photogrammetric bundle adjustment retrieves the unknown camera parameters and their theoretical accuracies. Examples, considerations and comparisons with real data and different case studies are illustrated to show the potentialities of the proposed methodology.

  15. The MINOS calibration detector

    This paper describes the MINOS calibration detector (CalDet) and the procedure used to calibrate it. The CalDet, a scaled-down but functionally equivalent model of the MINOS Far and Near detectors, was exposed to test beams in the CERN PS East Area during 2001-2003 to establish the response of the MINOS calorimeters to hadrons, electrons and muons in the range 0.2-10GeV/c. The CalDet measurements are used to fix the energy scale and constrain Monte Carlo simulations of MINOS

  16. Individual dosimetry and calibration

    In 1995 both the Individual Dosimetry and Calibration Sections worked under the condition of a status quo and concentrated fully on the routine part of their work. Nevertheless, the machine for printing the bar code which will be glued onto the film holder and hence identify the people when entering into high radiation areas was put into operation and most of the holders were equipped with the new identification. As far as the Calibration Section is concerned the project of the new source control system that is realized by the Technical Support Section was somewhat accelerated

  17. Calibration of farmer dosemeters

    The Farmer Dosemeters of Atomic Energy Medical Centre (AEMC) Jamshoro were calibrated in the Secondary Standard Dosimetry Laboratory (SSDL) at PINSTECH, using the NPL Secondary Standard Therapy level X-ray exposure meter. The results are presented in this report. (authors)

  18. Calibration Of Oxygen Monitors

    Zalenski, M. A.; Rowe, E. L.; Mcphee, J. R.


    Readings corrected for temperature, pressure, and humidity of air. Program for handheld computer developed to ensure accuracy of oxygen monitors in National Transonic Facility, where liquid nitrogen stored. Calibration values, determined daily, based on entries of data on barometric pressure, temperature, and relative humidity. Output provided directly in millivolts.

  19. Commodity-Free Calibration


    Commodity-free calibration is a reaction rate calibration technique that does not require the addition of any commodities. This technique is a specific form of the reaction rate technique, where all of the necessary reactants, other than the sample being analyzed, are either inherent in the analyzing system or specifically added or provided to the system for a reason other than calibration. After introduction, the component of interest is exposed to other reactants or flow paths already present in the system. The instrument detector records one of the following to determine the rate of reaction: the increase in the response of the reaction product, a decrease in the signal of the analyte response, or a decrease in the signal from the inherent reactant. With this data, the initial concentration of the analyte is calculated. This type of system can analyze and calibrate simultaneously, reduce the risk of false positives and exposure to toxic vapors, and improve accuracy. Moreover, having an excess of the reactant already present in the system eliminates the need to add commodities, which further reduces cost, logistic problems, and potential contamination. Also, the calculations involved can be simplified by comparison to those of the reaction rate technique. We conducted tests with hypergols as an initial investigation into the feasiblility of the technique.

  20. Calibration bench of flowmeters

    This equipment is devoted to the comparison of signals from two turbines installed in the Cabri experimental loop. The signal is compared to the standard turbine. The characteristics and the performance of the calibration bench are presented. (A.L.B.)

  1. Measurement System & Calibration report

    Vesth, Allan; Kock, Carsten Weber

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  2. Calibration with Absolute Shrinkage

    Øjelund, Henrik; Madsen, Henrik; Thyregod, Poul


    In this paper, penalized regression using the L-1 norm on the estimated parameters is proposed for chemometric je calibration. The algorithm is of the lasso type, introduced by Tibshirani in 1996 as a linear regression method with bound on the absolute length of the parameters, but a modification...

  3. Calibration issues for MUSE

    Kelz, Andreas; Roth, Martin; Bauer, Svend; Gerssen, Joris; Hahn, Thomas; Weilbacher, Peter; Laux, Uwe; Loupias, Magali; Kosmalski, Johan; McDermid, Richard; Bacon, Roland


    The Multi-Unit Spectroscopic Explorer (MUSE) is an integral-field spectrograph for the VLT for the next decade. Using an innovative field-splitting and slicing design, combined with an assembly of 24 spectrographs, MUSE will provide some 90,000 spectra in one exposure, which cover a simultaneous spectral range from 465 to 930nm. The design and manufacture of the Calibration Unit, the alignment tests of the Spectrograph and Detector sub-systems, and the development of the Data Reduction Software for MUSE are work-packages under the responsibility of the AIP, who is a partner in a European-wide consortium of 6 institutes and ESO, that is led by the Centre de Recherche Astronomique de Lyon. MUSE will be operated and therefore has to be calibrated in a variety of modes, which include seeing-limited and AO-assisted operations, providing a wide and narrow-field-of-view. MUSE aims to obtain unprecedented ultra-deep 3D-spectroscopic exposures, involving integration times of the order of 80 hours at the VLT. To achieve the corresponding science goals, instrumental stability, accurate calibration and adequate data reduction tools are needed. The paper describes the status at PDR of the AIP related work-packages, in particular with respect to the spatial, spectral, image quality, and geometrical calibration and related data reduction aspects.

  4. Entropic calibration revisited

    Brody, Dorje C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)]. E-mail:; Buckley, Ian R.C. [Centre for Quantitative Finance, Imperial College, London SW7 2AZ (United Kingdom); Constantinou, Irene C. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom); Meister, Bernhard K. [Blackett Laboratory, Imperial College, London SW7 2BZ (United Kingdom)


    The entropic calibration of the risk-neutral density function is effective in recovering the strike dependence of options, but encounters difficulties in determining the relevant greeks. By use of put-call reversal we apply the entropic method to the time reversed economy, which allows us to obtain the spot price dependence of options and the relevant greeks.

  5. Physiotherapy ultrasound calibrations

    Calibration of physiotherapy ultrasound equipment has long been a problem. Numerous surveys around the world over the past 20 years have all found that only a low percentage of the units tested had an output within 30% of that indicatd. In New Zealand, a survey carried out by the NRL in 1985 found that only 24% had an output, at the maximum setting, within + or - 20% of that indicated. The present performance Standard for new equipment (NZS 3200.2.5:1992) requires that the measured output should not deviate from that indicated by more than + or - 30 %. This may be tightened to + or - 20% in the next few years. Any calibration is only as good as the calibration equipment. Some force balances can be tested with small weights to simulate the force exerted by an ultrasound beam, but with others this is not possible. For such balances, testing may only be feasible with a calibrated source which could be used like a transfer standard. (author). 4 refs., 3 figs

  6. NVLAP calibration laboratory program

    Cigler, J.L.


    This paper presents an overview of the progress up to April 1993 in the development of the Calibration Laboratories Accreditation Program within the framework of the National Voluntary Laboratory Accreditation Program (NVLAP) at the National Institute of Standards and Technology (NIST).

  7. LOFAR Facet Calibration

    van Weeren, R. J.; Williams, W. L.; Hardcastle, M. J.; Shimwell, T. W.; Rafferty, D. A.; Sabater, J.; Heald, G.; Sridhar, S. S.; Dijkema, T. J.; Brunetti, G.; Brüggen, M.; Andrade-Santos, F.; Ogrean, G. A.; Röttgering, H. J. A.; Dawson, W. A.; Forman, W. R.; de Gasperin, F.; Jones, C.; Miley, G. K.; Rudnick, L.; Sarazin, C. L.; Bonafede, A.; Best, P. N.; Bîrzan, L.; Cassano, R.; Chyży, K. T.; Croston, J. H.; Ensslin, T.; Ferrari, C.; Hoeft, M.; Horellou, C.; Jarvis, M. J.; Kraft, R. P.; Mevius, M.; Intema, H. T.; Murray, S. S.; Orrú, E.; Pizzo, R.; Simionescu, A.; Stroe, A.; van der Tol, S.; White, G. J.


    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction-dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction-dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at ∼ 5\\prime\\prime resolution, meeting the specifications of the LOFAR Tier-1 northern survey.

  8. Radiation monitor calibration technique

    Reference radiations in the Secondary Standard Dosimetry Laboratory, OAEP have been improved and modified by employing lead attenuators. To identify low-level exposure rate, shadow-cone method has been applied. The secondary standard dosemeter has been used periodically to check the constancy of reference radiations to assure the calibration of dosemeters and dose-ratemeters used for radiation protection

  9. LOFAR facet calibration

    van Weeren, R J; Hardcastle, M J; Shimwell, T W; Rafferty, D A; Sabater, J; Heald, G; Sridhar, S S; Dijkema, T J; Brunetti, G; Brüggen, M; Andrade-Santos, F; Ogrean, G A; Röttgering, H J A; Dawson, W A; Forman, W R; de Gasperin, F; Jones, C; Miley, G K; Rudnick, L; Sarazin, C L; Bonafede, A; Best, P N; Bîrzan, L; Cassano, R; Chyży, K T; Croston, J H; Ensslin, T; Ferrari, C; Hoeft, M; Horellou, C; Jarvis, M J; Kraft, R P; Mevius, M; Intema, H T; Murray, S S; Orrú, E; Pizzo, R; Simionescu, A; Stroe, A; van der Tol, S; White, G J


    LOFAR, the Low-Frequency Array, is a powerful new radio telescope operating between 10 and 240 MHz. LOFAR allows detailed sensitive high-resolution studies of the low-frequency radio sky. At the same time LOFAR also provides excellent short baseline coverage to map diffuse extended emission. However, producing high-quality deep images is challenging due to the presence of direction dependent calibration errors, caused by imperfect knowledge of the station beam shapes and the ionosphere. Furthermore, the large data volume and presence of station clock errors present additional difficulties. In this paper we present a new calibration scheme, which we name facet calibration, to obtain deep high-resolution LOFAR High Band Antenna images using the Dutch part of the array. This scheme solves and corrects the direction dependent errors in a number of facets that cover the observed field of view. Facet calibration provides close to thermal noise limited images for a typical 8 hr observing run at $\\sim$ 5arcsec resolu...

  10. Pseudo Linear Gyro Calibration

    Harman, Richard; Bar-Itzhack, Itzhack Y.


    Previous high fidelity onboard attitude algorithms estimated only the spacecraft attitude and gyro bias. The desire to promote spacecraft and ground autonomy and improvements in onboard computing power has spurred development of more sophisticated calibration algorithms. Namely, there is a desire to provide for sensor calibration through calibration parameter estimation onboard the spacecraft as well as autonomous estimation on the ground. Gyro calibration is a particularly challenging area of research. There are a variety of gyro devices available for any prospective mission ranging from inexpensive low fidelity gyros with potentially unstable scale factors to much more expensive extremely stable high fidelity units. Much research has been devoted to designing dedicated estimators such as particular Extended Kalman Filter (EKF) algorithms or Square Root Information Filters. This paper builds upon previous attitude, rate, and specialized gyro parameter estimation work performed with Pseudo Linear Kalman Filter (PSELIKA). The PSELIKA advantage is the use of the standard linear Kalman Filter algorithm. A PSELIKA algorithm for an orthogonal gyro set which includes estimates of attitude, rate, gyro misalignments, gyro scale factors, and gyro bias is developed and tested using simulated and flight data. The measurements PSELIKA uses include gyro and quaternion tracker data.

  11. Pleiades Absolute Calibration : Inflight Calibration Sites and Methodology

    Lachérade, S.; Fourest, S.; Gamet, P.; Lebègue, L.


    In-flight calibration of space sensors once in orbit is a decisive step to be able to fulfil the mission objectives. This article presents the methods of the in-flight absolute calibration processed during the commissioning phase. Four In-flight calibration methods are used: absolute calibration, cross-calibration with reference sensors such as PARASOL or MERIS, multi-temporal monitoring and inter-bands calibration. These algorithms are based on acquisitions over natural targets such as African deserts, Antarctic sites, La Crau (Automatic calibration station) and Oceans (Calibration over molecular scattering) or also new extra-terrestrial sites such as the Moon and selected stars. After an overview of the instrument and a description of the calibration sites, it is pointed out how each method is able to address one or several aspects of the calibration. We focus on how these methods complete each other in their operational use, and how they help building a coherent set of information that addresses all aspects of in-orbit calibration. Finally, we present the perspectives that the high level of agility of PLEIADES offers for the improvement of its calibration and a better characterization of the calibration sites.

  12. Mercury CEM Calibration

    John F. Schabron; Joseph F. Rovani; Susan S. Sorini


    The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005, requires that calibration of mercury continuous emissions monitors (CEMs) be performed with NIST-traceable standards. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor generators. The traceability protocol will be written by EPA. Traceability will be based on the actual analysis of the output of each calibration unit at several concentration levels ranging from about 2-40 ug/m{sup 3}, and this analysis will be directly traceable to analyses by NIST using isotope dilution inductively coupled plasma/mass spectrometry (ID ICP/MS) through a chain of analyses linking the calibration unit in the power plant to the NIST ID ICP/MS. Prior to this project, NIST did not provide a recommended mercury vapor pressure equation or list mercury vapor pressure in its vapor pressure database. The NIST Physical and Chemical Properties Division in Boulder, Colorado was subcontracted under this project to study the issue in detail and to recommend a mercury vapor pressure equation that the vendors of mercury vapor pressure calibration units can use to calculate the elemental mercury vapor concentration in an equilibrium chamber at a particular temperature. As part of this study, a preliminary evaluation of calibration units from five vendors was made. The work was performed by NIST in Gaithersburg, MD and Joe Rovani from WRI who traveled to NIST as a Visiting Scientist.

  13. Calibrated Properties Model

    The purpose of this Model Report is to document the Calibrated Properties Model that provides calibrated parameter sets for unsaturated zone (UZ) flow and transport process models for the Office of Repository Development (ORD). The UZ contains the unsaturated rock layers overlying the repository and host unit, which constitute a natural barrier to flow, and the unsaturated rock layers below the repository which constitute a natural barrier to flow and transport. This work followed, and was planned in, ''Technical Work Plan (TWP) for: Performance Assessment Unsaturated Zone'' (BSC 2002 [160819], Section 1.10.8 [under Work Package (WP) AUZM06, Climate Infiltration and Flow], and Section I-1-1 [in Attachment I, Model Validation Plans]). In Section 4.2, four acceptance criteria (ACs) are identified for acceptance of this Model Report; only one of these (Section, AC 3) was identified in the TWP (BSC 2002 [160819], Table 3-1). These calibrated property sets include matrix and fracture parameters for the UZ Flow and Transport Model (UZ Model), drift seepage models, and drift-scale and mountain-scale coupled-process models from the UZ Flow, Transport and Coupled Processes Department in the Natural Systems Subproject of the Performance Assessment (PA) Project. The Calibrated Properties Model output will also be used by the Engineered Barrier System Department in the Engineering Systems Subproject. The Calibrated Properties Model provides input through the UZ Model and other process models of natural and engineered systems to the Total System Performance Assessment (TSPA) models, in accord with the PA Strategy and Scope in the PA Project of the Bechtel SAIC Company, LLC (BSC). The UZ process models provide the necessary framework to test conceptual hypotheses of flow and transport at different scales and predict flow and transport behavior under a variety of climatic and thermal-loading conditions. UZ flow is a TSPA model component

  14. Field calibration of cup anemometers

    Schmidt Paulsen, Uwe; Mortensen, Niels Gylling; Hansen, Jens Carsten;


    A field calibration method and results are described along with the experience gained with the method. The cup anemometers to be calibrated are mounted in a row on a 10-m high rig and calibrated in the free wind against a reference cup anemometer. The method has been reported [1] to improve the...... statistical bias on the data relative to calibrations carried out in a wind tunnel. The methodology is sufficiently accurate for calibration of cup anemometers used for wind resource assessments and provides a simple, reliable and cost-effective solution to cup anemometer calibration, especially suited for...

  15. Mercury Calibration System

    John Schabron; Eric Kalberer; Joseph Rovani; Mark Sanderson; Ryan Boysen; William Schuster


    U.S. Environmental Protection Agency (EPA) Performance Specification 12 in the Clean Air Mercury Rule (CAMR) states that a mercury CEM must be calibrated with National Institute for Standards and Technology (NIST)-traceable standards. In early 2009, a NIST traceable standard for elemental mercury CEM calibration still does not exist. Despite the vacature of CAMR by a Federal appeals court in early 2008, a NIST traceable standard is still needed for whatever regulation is implemented in the future. Thermo Fisher is a major vendor providing complete integrated mercury continuous emissions monitoring (CEM) systems to the industry. WRI is participating with EPA, EPRI, NIST, and Thermo Fisher towards the development of the criteria that will be used in the traceability protocols to be issued by EPA. An initial draft of an elemental mercury calibration traceability protocol was distributed for comment to the participating research groups and vendors on a limited basis in early May 2007. In August 2007, EPA issued an interim traceability protocol for elemental mercury calibrators. Various working drafts of the new interim traceability protocols were distributed in late 2008 and early 2009 to participants in the Mercury Standards Working Committee project. The protocols include sections on qualification and certification. The qualification section describes in general terms tests that must be conducted by the calibrator vendors to demonstrate that their calibration equipment meets the minimum requirements to be established by EPA for use in CAMR monitoring. Variables to be examined include linearity, ambient temperature, back pressure, ambient pressure, line voltage, and effects of shipping. None of the procedures were described in detail in the draft interim documents; however they describe what EPA would like to eventually develop. WRI is providing the data and results to EPA for use in developing revised experimental procedures and realistic acceptance criteria based on

  16. Streak camera time calibration procedures

    Long, J.; Jackson, I.


    Time calibration procedures for streak cameras utilizing a modulated laser beam are described. The time calibration determines a writing rate accuracy of 0.15% with a rotating mirror camera and 0.3% with an image converter camera.

  17. The Calibration Reference Data System

    Greenfield, P.; Miller, T.


    We describe a software architecture and implementation for using rules to determine which calibration files are appropriate for calibrating a given observation. This new system, the Calibration Reference Data System (CRDS), replaces what had been previously used for the Hubble Space Telescope (HST) calibration pipelines, the Calibration Database System (CDBS). CRDS will be used for the James Webb Space Telescope (JWST) calibration pipelines, and is currently being used for HST calibration pipelines. CRDS can be easily generalized for use in similar applications that need a rules-based system for selecting the appropriate item for a given dataset; we give some examples of such generalizations that will likely be used for JWST. The core functionality of the Calibration Reference Data System is available under an Open Source license. CRDS is briefly contrasted with a sampling of other similar systems used at other observatories.

  18. Optical tweezers absolute calibration

    Dutra, R S; Neto, P A Maia; Nussenzveig, H M


    Optical tweezers are highly versatile laser traps for neutral microparticles, with fundamental applications in physics and in single molecule cell biology. Force measurements are performed by converting the stiffness response to displacement of trapped transparent microspheres, employed as force transducers. Usually, calibration is indirect, by comparison with fluid drag forces. This can lead to discrepancies by sizable factors. Progress achieved in a program aiming at absolute calibration, conducted over the past fifteen years, is briefly reviewed. Here we overcome its last major obstacle, a theoretical overestimation of the peak stiffness, within the most employed range for applications, and we perform experimental validation. The discrepancy is traced to the effect of primary aberrations of the optical system, which are now included in the theory. All required experimental parameters are readily accessible. Astigmatism, the dominant effect, is measured by analyzing reflected images of the focused laser spo...

  19. Astrid-2 SSC ASUMagnetic Calibration

    Primdahl, Fritz


    Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory.......Report of the inter calibration between the starcamera and the fluxgate magnetometer onboard the ASTRID-2 satellite. This calibration was performed in the night between the 15. and 16. May 1997 at the Lovö magnetic observatory....

  20. Optical Calibration of SNO+

    Maneira, J.; Peeters, S.; Sinclair, J.


    SNO is being upgraded to SNO+, which has as its main goal the search for neutrinoless double-beta decay. The upgrade is defined by filling with a novel scintillator mixture containing 130Te. With a lower energy threshold than SNO, SNO+ will be sensitive to other exciting new physics. Here we are describing new optical calibration system that meets new, more stringent radiopurity requirements has been developed.

  1. Camera Calibration Using Silhouettes

    Boyer, Edmond


    This report addresses the problem of estimating camera parameters from images where object silhouettes only are known. Several modeling applications make use of silhouettes, and while calibration methods are well known when considering points or lines matched along image sequences, the problem appears to be more difficult when considering silhouettes. However, such primitives encode also information on camera parameters by the fact that their associated viewing cones should present a common i...

  2. Program Calibrates Strain Gauges

    Okazaki, Gary D.


    Program dramatically reduces personnel and time requirements for acceptance tests of hardware. Data-acquisition system reads output from Wheatstone full-bridge strain-gauge circuit and calculates strain by use of shunt calibration technique. Program nearly instantaneously tabulates and plots strain data against load-cell outputs. Modified to acquire strain data for other specimens wherever full-bridge strain-gauge circuits used. Written in HP BASIC.

  3. Calibration specimens for microscopy

    Kolařík, Vladimír; Matějka, Milan; Matějka, František; Krátký, Stanislav; Urbánek, Michal; Horáček, Miroslav; Král, Stanislav; Bok, Jan

    Ostrava: TANGER Ltd, 2012, s. 713-716. ISBN 978-80-87294-32-1. [NANOCON 2012. International Conference /4./. Brno (CZ), 23.10.2012-25.10.2012] R&D Projects: GA MŠk ED0017/01/01; GA TA ČR TE01020233; GA MPO FR-TI1/576 Institutional support: RVO:68081731 Keywords : E-beam technology * calibration specimen * scanning electron microscopy Subject RIV: JA - Electronics ; Optoelectronics, Electrical Engineering

  4. Calibrating bacterial evolution

    Ochman, Howard; Elwyn, Susannah; Moran, Nancy A


    Attempts to calibrate bacterial evolution have relied on the assumption that rates of molecular sequence divergence in bacteria are similar to those of higher eukaryotes, or to those of the few bacterial taxa for which ancestors can be reliably dated from ecological or geological evidence. Despite similarities in the substitution rates estimated for some lineages, comparisons of the relative rates of evolution at different classes of nucleotide sites indicate no basis for their universal appl...

  5. Mesoscale hybrid calibration artifact

    Tran, Hy D.; Claudet, Andre A.; Oliver, Andrew D.


    A mesoscale calibration artifact, also called a hybrid artifact, suitable for hybrid dimensional measurement and the method for make the artifact. The hybrid artifact has structural characteristics that make it suitable for dimensional measurement in both vision-based systems and touch-probe-based systems. The hybrid artifact employs the intersection of bulk-micromachined planes to fabricate edges that are sharp to the nanometer level and intersecting planes with crystal-lattice-defined angles.

  6. Dosimetry and Calibration Section

    The two tasks of the Dosimetry and Calibration Section at CERN are the Individual Dosimetry Service which assures the personal monitoring of about 5000 persons potentially exposed to ionizing radiation at CERN, and the Calibration Laboratory which verifies all the instruments and monitors. This equipment is used by the sections of the RP Group for assuring radiation protection around CERN's accelerators, and by the Environmental Section of TISTE. In addition, nearly 250 electronic and 300 quartz fibre dosimeters, employed in operational dosimetry, are calibrated at least once a year. The Individual Dosimetry Service uses an extended database (INDOS) which contains information about all the individual doses ever received at CERN. For most of 1997 it was operated without the support of a database administrator as the technician who had assured this work retired. The Software Support Section of TIS-TE took over the technical responsibility of the database, but in view of the many other tasks of this Section and the lack of personnel, only a few interventions for solving immediate problems were possible

  7. Calibrated Properties Model

    The purpose of this model report is to document the calibrated properties model that provides calibrated property sets for unsaturated zone (UZ) flow and transport process models (UZ models). The calibration of the property sets is performed through inverse modeling. This work followed, and was planned in, ''Technical Work Plan (TWP) for: Unsaturated Zone Flow Analysis and Model Report Integration'' (BSC 2004 [DIRS 169654], Sections 1.2.6 and Direct inputs to this model report were derived from the following upstream analysis and model reports: ''Analysis of Hydrologic Properties Data'' (BSC 2004 [DIRS 170038]); ''Development of Numerical Grids for UZ Flow and Transport Modeling'' (BSC 2004 [DIRS 169855]); ''Simulation of Net Infiltration for Present-Day and Potential Future Climates'' (BSC 2004 [DIRS 170007]); ''Geologic Framework Model'' (GFM2000) (BSC 2004 [DIRS 170029]). Additionally, this model report incorporates errata of the previous version and closure of the Key Technical Issue agreement TSPAI 3.26 (Section 6.2.2 and Appendix B), and it is revised for improved transparency

  8. Del Piero

    Umbaca, Enzo


    Rifacendosi al “Pinturicchio”, epiteto con cui Gianni Agnelli consacrava il calciatore Alessandro Del Piero un “artista del calcio”, Enzo Umbaca scrisse al popolare personaggio invitandolo a collaborare ad una performance durante la quale Pinturicchio-Del Piero avrebbe affrescato il muro di una galleria torinese calciando un pallone macchiato di grafite contro un muro sul quale è appesa la copia di un opera del Pinturicchio.

  9. Calibration effects on orbit determination

    Madrid, G. A.; Winn, F. B.; Zielenbach, J. W.; Yip, K. B.


    The effects of charged particle and tropospheric calibrations on the orbit determination (OD) process are analyzed. The calibration process consisted of correcting the Doppler observables for the media effects. Calibrated and uncalibrated Doppler data sets were used to obtain OD results for past missions as well as Mariner Mars 1971. Comparisons of these Doppler reductions show the significance of the calibrations. For the MM'71 mission, the media calibrations proved themselves effective in diminishing the overall B-plane error and reducing the Doppler residual signatures.

  10. Mammography calibration: Factor or fit?

    Dose measurements in mammography x-ray have become more important and a basic path in quality assurance programmes. It is recognized by the international guidelines that it is necessary to have calibration services offered for mammography beams in order to help the improvement of the clinical diagnosis. Major efforts have been made by several laboratories in order to establish an appropriate and traceable calibration infrastructure and to provide the basis for a quality control programme in mammography. The indication of a dosimeter, whose reference point is positioned at the point of test, is compared with the conventional true value of the quantity to be measured. The calibration coefficient is then the ratio of the conventional true value to the indicated. The Primary Standard Dosimetry Laboratory - PSDL or the Secondary Standard Dosimetry Laboratory - SSDL provides the calibration coefficient of the dosimeters in reference to the Half Value Layers - HVL implemented in their laboratories. The dosimetry calibration data is enough when the user has the same system as the laboratory where the ionization chamber has been calibrated. However, there are other calibration systems that have different calibration qualities implemented using different combinations of anode and filter and, therefore, there is no direct relation with the calibration coefficient. How to deal with this? There are two different ways to obtain calibration coefficients when the user's implemented qualities are different from the calibration laboratory's qualities. The first is the interpolation of each calibration coefficient stated in the certificate. The second is the fit of all calibration coefficients, separately for non-attenuated and attenuated beam qualities, to obtain a function by which the calibration coefficients can be determined at each beam quality. The second one includes the statistical fluctuation. The dosimetry calibration data must fit an analytical form, as for example a

  11. A variable acceleration calibration system

    Johnson, Thomas H.


    A variable acceleration calibration system that applies loads using gravitational and centripetal acceleration serves as an alternative, efficient and cost effective method for calibrating internal wind tunnel force balances. Two proof-of-concept variable acceleration calibration systems are designed, fabricated and tested. The NASA UT-36 force balance served as the test balance for the calibration experiments. The variable acceleration calibration systems are shown to be capable of performing three component calibration experiments with an approximate applied load error on the order of 1% of the full scale calibration loads. Sources of error are indentified using experimental design methods and a propagation of uncertainty analysis. Three types of uncertainty are indentified for the systems and are attributed to prediction error, calibration error and pure error. Angular velocity uncertainty is shown to be the largest indentified source of prediction error. The calibration uncertainties using a production variable acceleration based system are shown to be potentially equivalent to current methods. The production quality system can be realized using lighter materials and a more precise instrumentation. Further research is needed to account for balance deflection, forcing effects due to vibration, and large tare loads. A gyroscope measurement technique is shown to be capable of resolving the balance deflection angle calculation. Long term research objectives include a demonstration of a six degree of freedom calibration, and a large capacity balance calibration.

  12. Automatic force balance calibration system

    Ferris, Alice T.


    A system for automatically calibrating force balances is provided. The invention uses a reference balance aligned with the balance being calibrated to provide superior accuracy while minimizing the time required to complete the calibration. The reference balance and the test balance are rigidly attached together with closely aligned moment centers. Loads placed on the system equally effect each balance, and the differences in the readings of the two balances can be used to generate the calibration matrix for the test balance. Since the accuracy of the test calibration is determined by the accuracy of the reference balance and current technology allows for reference balances to be calibrated to within +/-0.05% the entire system has an accuracy of +/-0.2%. The entire apparatus is relatively small and can be mounted on a movable base for easy transport between test locations. The system can also accept a wide variety of reference balances, thus allowing calibration under diverse load and size requirements.

  13. Structured light camera calibration

    Garbat, P.; Skarbek, W.; Tomaszewski, M.


    Structured light camera which is being designed with the joined effort of Institute of Radioelectronics and Institute of Optoelectronics (both being large units of the Warsaw University of Technology within the Faculty of Electronics and Information Technology) combines various hardware and software contemporary technologies. In hardware it is integration of a high speed stripe projector and a stripe camera together with a standard high definition video camera. In software it is supported by sophisticated calibration techniques which enable development of advanced application such as real time 3D viewer of moving objects with the free viewpoint or 3D modeller for still objects.

  14. Smart Calibration of Excavators

    Bro, Marie; Døring, Kasper; Ellekilde, Lars-Peter; Hansen, Mikael Sonne; Markvorsen, Steen; Spence, David; Stolpe, Mathias; Sølvason, Dorthe


    Excavators dig holes. But where is the bucket? The purpose of this report is to treat four different problems concerning calibrations of position indicators for excavators in operation at concrete construction sites. All four problems are related to the question of how to determine the precise...... geographic and/or site-relative position of a given excavator and its bucket. However, our presentations and solutions to the problems can, nevertheless, be read and studied in any order and independently of each other. This also implies and induces a gentle warning to the reader: The {\\em{notation}} need...

  15. Use of Radiometrically Calibrated Flat-Plate Calibrators in Calibration of Radiation Thermometers

    Cárdenas-García, D.; Méndez-Lango, E.


    Most commonly used, low-temperature, infrared thermometers have large fields of view sizes that make them difficult to be calibrated with narrow aperture blackbodies. Flat-plate calibrators with large emitting surfaces have been proposed for calibrating these infrared thermometers. Because the emissivity of the flat plate is not unity, its radiance temperature is wavelength dependent. For calibration, the wavelength pass band of the device under test should match that of the reference infrared thermometer. If the device under test and reference radiometer have different pass bands, then it is possible to calculate the corresponding correction if the emissivity of the flat plate is known. For example, a correction of at is required when calibrating a infrared thermometer with a "" radiometrically calibrated flat-plate calibrator. A method is described for using a radiometrically calibrated flat-plate calibrator that covers both cases of match and mismatch working wavelength ranges of a reference infrared thermometer and infrared thermometers to be calibrated with the flat-plate calibrator. Also, an application example is included in this paper.

  16. Dosimetry and Calibration Section

    The Dosimetry and Calibration Section fulfils two tasks within CERN's Radiation Protection Group: the Individual Dosimetry Service monitors more than 5000 persons potentially exposed to ionizing radiation on the CERN sites, and the Calibration Laboratory verifies throughout the year, at regular intervals, over 1000 instruments, monitors, and electronic dosimeters used by RP Group. The establishment of a Quality Assurance System for the Individual Dosimetry Service, a requirement of the new Swiss Ordinance for personal dosimetry, put a considerable workload on the section. Together with an external consultant it was decided to identify and then describe the different 'processes' of the routine work performed in the dosimetry service. The resulting Quality Manual was submitted to the Federal Office for Public Health in Bern in autumn. The CERN Individual Dosimetry Service will eventually be officially endorsed after a successful technical test in March 1999. On the technical side, the introduction of an automatic development machine for gamma films was very successful. It processes the dosimetric films without an operator being present, and its built-in regeneration mechanism keeps the concentration of the processing chemicals at a constant level

  17. NIR prediction of fruit moisture, free acidity and oil content in intact olives


    Full Text Available In this work, the prediction of olive fruit and virgin olive oil quality parameters through the direct measuring of the fruit using near infrared spectrometry (NIRS has been investigated and the effectiveness of a portable spectrometer has been assessed. Models and calibration tests were developed using both the hexane-isopropanol extraction of individual olive fruits, and the Soxhlet extraction of olive paste. The parameters analyzed were the free acidity in olive oil, oil yield from physical extraction, oil content referring to fresh weight, oil content referring to dry matter and fruit moisture. The results indicate a good predictive potential with both methodologies and serve to encourage improvement in the obtained models through the enlargement of the calibrations. Fruit moisture prediction models showed high accuracy.

    En este trabajo se ha investigado la predicción de parámetros de calidad de aceitunas y de aceite de oliva virgen mediante medidas directas en el fruto de espectrometría de infrarrojo cercano (NIRS, evaluándose la utilidad de un espectrómetro portátil. Se han desarrollado respectivamente modelos predictivos y calibraciones utilizando como análisis de referencia tanto la extracción de aceitunas individualmente con hexano-isopropanol, como la extracción de pasta de aceituna mediante Soxhlet. Los parámetros analizados fueron: acidez libre del aceite, rendimiento de la extracción física de aceite, contenido de aceite referido a peso fresco, contenido de aceite referido a materia seca y humedad del fruto. Los resultados indican un buen potencial de predicción mediante ambos métodos y alientan al perfeccionamiento de los modelos obtenidos mediante la ampliación de las calibraciones. Los modelos predictivos de la humedad del fruto mostraron una alta precisión.

  18. The Detector Calibration System for the CUORE cryogenic bolometer array

    Cushman, J S; Davis, C J; Ejzak, L; Lenz, D; Lim, K E; Heeger, K M; Maruyama, R H; Nucciotti, A; Sangiorgio, S; Wise, T


    The Cryogenic Underground Observatory for Rare Events (CUORE) is a ton-scale cryogenic experiment designed to search for neutrinoless double-beta decay of $^{130}$Te and other rare events. The CUORE detector consists of 988 TeO$_2$ bolometers operated underground at 10~mK in a dilution refrigerator at the Laboratori Nazionali del Gran Sasso. Candidate events are identified through a precise measurement of their energy. The absolute energy response of the detectors is established by the regular calibration of each individual bolometer using gamma sources. The close-packed configuration of the CUORE bolometer array combined with the extensive shielding surrounding the detectors requires the placement of calibration sources within the array itself. The CUORE Detector Calibration System is designed to insert radioactive sources into and remove them from the cryostat while respecting the stringent heat load, radiopurity, and operational requirements of the experiment. This paper describes the design, commissioning...

  19. Camera Calibration: a USU Implementation

    Ma, Lili; Chen, YangQuan; Moore, Kevin L.


    The task of camera calibration is to estimate the intrinsic and extrinsic parameters of a camera model. Though there are some restricted techniques to infer the 3-D information about the scene from uncalibrated cameras, effective camera calibration procedures will open up the possibility of using a wide range of existing algorithms for 3-D reconstruction and recognition. The applications of camera calibration include vision-based metrology, robust visual platooning and visual docking of mobil...

  20. Calibration procedure for zenith plummets

    Jelena GUČEVIĆ


    Full Text Available Zenith plummets are used mainly in applied geodesy, in civil engineering surveying, for materialization of the local vertical. The error of the vertical deflection of the instrument is directly transferred to the error of the observing construction. That is why a proper calibration procedure for the zenithlot is required. Metrological laboratory of the Faculty of Civil Engineering in Belgrade developed such calibration procedure. Here we present a mathematical model of the calibration and some selected results.

  1. Calibration procedure for zenith plummets

    Jelena GUČEVIĆ; Delčev, Siniša; Vukan OGRIZOVIĆ


    Zenith plummets are used mainly in applied geodesy, in civil engineering surveying, for materialization of the local vertical. The error of the vertical deflection of the instrument is directly transferred to the error of the observing construction. That is why a proper calibration procedure for the zenithlot is required. Metrological laboratory of the Faculty of Civil Engineering in Belgrade developed such calibration procedure. Here we present a mathematical model of the calibration and som...

  2. Radiological Calibration and Standards Facility

    Federal Laboratory Consortium — PNNL maintains a state-of-the-art Radiological Calibration and Standards Laboratory on the Hanford Site at Richland, Washington. Laboratory staff provide expertise...

  3. Calibration Techniques for VERITAS

    Hanna, David


    VERITAS is an array of four identical telescopes designed for detecting and measuring astrophysical gamma rays with energies in excess of 100 GeV. Each telescope uses a 12 m diameter reflector to collect Cherenkov light from air showers initiated by incident gamma rays and direct it onto a `camera' comprising 499 photomultiplier tubes read out by flash ADCs. We describe here calibration methods used for determining the values of the parameters which are necessary for converting the digitized PMT pulses to gamma-ray energies and directions. Use of laser pulses to determine and monitor PMT gains is discussed, as are measurements of the absolute throughput of the telescopes using muon rings.

  4. TOD to TTP calibration

    Bijl, Piet; Reynolds, Joseph P.; Vos, Wouter K.; Hogervorst, Maarten A.; Fanning, Jonathan D.


    The TTP (Targeting Task Performance) metric, developed at NVESD, is the current standard US Army model to predict EO/IR Target Acquisition performance. This model however does not have a corresponding lab or field test to empirically assess the performance of a camera system. The TOD (Triangle Orientation Discrimination) method, developed at TNO in The Netherlands, provides such a measurement. In this study, we make a direct comparison between TOD performance for a range of sensors and the extensive historical US observer performance database built to develop and calibrate the TTP metric. The US perception data were collected doing an identification task by military personnel on a standard 12 target, 12 aspect tactical vehicle image set that was processed through simulated sensors for which the most fundamental sensor parameters such as blur, sampling, spatial and temporal noise were varied. In the present study, we measured TOD sensor performance using exactly the same sensors processing a set of TOD triangle test patterns. The study shows that good overall agreement is obtained when the ratio between target characteristic size and TOD test pattern size at threshold equals 6.3. Note that this number is purely based on empirical data without any intermediate modeling. The calibration of the TOD to the TTP is highly beneficial to the sensor modeling and testing community for a variety of reasons. These include: i) a connection between requirement specification and acceptance testing, and ii) a very efficient method to quickly validate or extend the TTP range prediction model to new systems and tasks.

  5. Tectonic calibrations in molecular dating



    Molecular dating techniques require the use of calibrations, which are usually fossil or geological vicariance-based.Fossil calibrations have been criticised because they result only in minimum age estimates. Based on a historical biogeographic perspective, Ⅰ suggest that vicariance-based calibrations are more dangerous. Almost all analytical methods in historical biogeography are strongly biased towards inferring vicariance, hence vicariance identified through such methods is unreliable. Other studies, especially of groups found on Gondwanan fragments, have simply assumed vicariance. Although it was previously believed that vicariance was the predominant mode of speciation, mounting evidence now indicates that speciation by dispersal is common, dominating vicariance in several groups. Moreover, the possibility of speciation having occurred before the said geological event cannot be precluded. Thus, geological calibrations can under- or overestimate times, whereas fossil calibrations always result in minimum estimates. Another major drawback of vicariant calibrations is the problem of circular reasoning when the resulting estimates are used to infer ages of biogeographic events. Ⅰ argue that fossil-based dating is a superior alternative to vicariance, primarily because the strongest assumption in the latter, that speciation was caused by the said geological process, is more often than not the most tenuous. When authors prefer to use a combination of fossil and vicariant calibrations, one suggestion is to report results both with and without inclusion of the geological constraints. Relying solely on vicariant calibrations should be strictly avoided.

  6. Field calibration of cup anemometers

    Kristensen, L.; Jensen, G.; Hansen, A.;


    An outdoor calibration facility for cup anemometers, where the signals from 10 anemometers of which at least one is a reference can be can be recorded simultaneously, has been established. The results are discussed with special emphasis on the statisticalsignificance of the calibration expressions...

  7. Liquid Krypton Calorimeter Calibration Software

    Hughes, Christina Lindsay


    Calibration of the liquid krypton calorimeter (LKr) of the NA62 experiment is managed by a set of standalone programs, or an online calibration driver. These programs are similar to those used by NA48, but have been updated to utilize classes and translated to C++ while maintaining a common functionality. A set of classes developed to handle communication with hardware was used to develop the three standalone programs as well as the main driver program for online calibration between bursts. The main calibration driver has been designed to respond to run control commands and receive burst data, both transmitted via DIM. In order to facilitate the process of reading in calibration parameters, a serializable class has been introduced, allowing the replacement of standard text files with XML configuration files.

  8. The Advanced LIGO Photon Calibrators

    Karki, S; Kandhasamy, S; Abbott, B P; Abbott, T D; Anders, E H; Berliner, J; Betzwieser, J; Daveloza, H P; Cahillane, C; Canete, L; Conley, C; Gleason, J R; Goetz, E; Kissel, J S; Izumi, K; Mendell, G; Quetschke, V; Rodruck, M; Sachdev, S; Sadecki, T; Schwinberg, P B; Sottile, A; Wade, M; Weinstein, A J; West, M; Savage, R L


    The two interferometers of the Laser Interferometry Gravitaional-wave Observatory (LIGO) recently detected gravitational waves from the mergers of binary black hole systems. Accurate calibration of the output of these detectors was crucial for the observation of these events, and the extraction of parameters of the sources. The principal tools used to calibrate the responses of the second-generation (Advanced) LIGO detectors to gravitational waves are systems based on radiation pressure and referred to as Photon Calibrators. These systems, which were completely redesigned for Advanced LIGO, include several significant upgrades that enable them to meet the calibration requirements of second-generation gravitational wave detectors in the new era of gravitational-wave astronomy. We report on the design, implementation, and operation of these Advanced LIGO Photon Calibrators that are currently providing fiducial displacements on the order of $10^{-18}$ m/$\\sqrt{\\textrm{Hz}}$ with accuracy and precision of better ...

  9. Antenna Calibration and Measurement Equipment

    Rochblatt, David J.; Cortes, Manuel Vazquez


    A document describes the Antenna Calibration & Measurement Equipment (ACME) system that will provide the Deep Space Network (DSN) with instrumentation enabling a trained RF engineer at each complex to perform antenna calibration measurements and to generate antenna calibration data. This data includes continuous-scan auto-bore-based data acquisition with all-sky data gathering in support of 4th order pointing model generation requirements. Other data includes antenna subreflector focus, system noise temperature and tipping curves, antenna efficiency, reports system linearity, and instrument calibration. The ACME system design is based on the on-the-fly (OTF) mapping technique and architecture. ACME has contributed to the improved RF performance of the DSN by approximately a factor of two. It improved the pointing performances of the DSN antennas and productivity of its personnel and calibration engineers.


    Smith, V.L.; Carstensen, H.K.


    An improved time calibrated sweep circuit is presented, which extends the range of usefulness of conventional oscilloscopes as utilized for time calibrated display applications in accordance with U. S. Patent No. 2,832,002. Principal novelty resides in the provision of a pair of separate signal paths, each of which is phase and amplitude adjustable, to connect a high-frequency calibration oscillator to the output of a sawtooth generator also connected to the respective horizontal deflection plates of an oscilloscope cathode ray tube. The amplitude and phase of the calibration oscillator signals in the two signal paths are adjusted to balance out feedthrough currents capacitively coupled at high frequencies of the calibration oscillator from each horizontal deflection plate to the vertical plates of the cathode ray tube.

  11. Requirements for gamma radiation survey meter calibration

    This guide describes the minimum requirements for calibrating a portable analog gamma radiation survey meter by means of a beam calibrator, with a known calibration source. If an alternative method of calibration is to be used the licensee should make a written request to the Atomic Energy Control Board that describes the calibration method to be used, and request the Board's permission to use that method in place of the requirements contained in this guide. This guide explains: the responsibility for survey meter calibration if licensees calibrate their own survey meters, use the services of a Canadian calibration agency, and use the services of a non-Canadian calibration agency; the requirements for survey meter calibration and the supporting documentation; the requirements for record-keeping; and, a calibration certificate, a calibration sticker, and a notification of failure to calibrate form, with examples

  12. Mercury Continuous Emmission Monitor Calibration

    John Schabron; Eric Kalberer; Ryan Boysen; William Schuster; Joseph Rovani


    Mercury continuous emissions monitoring systems (CEMs) are being implemented in over 800 coal-fired power plant stacks throughput the U.S. Western Research Institute (WRI) is working closely with the Electric Power Research Institute (EPRI), the National Institute of Standards and Technology (NIST), and the Environmental Protection Agency (EPA) to facilitate the development of the experimental criteria for a NIST traceability protocol for dynamic elemental mercury vapor calibrators/generators. These devices are used to calibrate mercury CEMs at power plant sites. The Clean Air Mercury Rule (CAMR) which was published in the Federal Register on May 18, 2005 and vacated by a Federal appeals court in early 2008 required that calibration be performed with NIST-traceable standards. Despite the vacature, mercury emissions regulations in the future will require NIST traceable calibration standards, and EPA does not want to interrupt the effort towards developing NIST traceability protocols. The traceability procedures will be defined by EPA. An initial draft traceability protocol was issued by EPA in May 2007 for comment. In August 2007, EPA issued a conceptual interim traceability protocol for elemental mercury calibrators. The protocol is based on the actual analysis of the output of each calibration unit at several concentration levels ranging initially from about 2-40 {micro}g/m{sup 3} elemental mercury, and in the future down to 0.2 {micro}g/m{sup 3}, and this analysis will be directly traceable to analyses by NIST. The EPA traceability protocol document is divided into two separate sections. The first deals with the qualification of calibrator models by the vendors for use in mercury CEM calibration. The second describes the procedure that the vendors must use to certify the calibrators that meet the qualification specifications. The NIST traceable certification is performance based, traceable to analysis using isotope dilution inductively coupled plasma

  13. del Nim

    Arturo Martínez Rodríguez


    Full Text Available Dado su uso específico como fuente de insecticidas naturales de fácil biodegradación, entre otras ventajas, el cultivo del árbol Nim ha sido mundialmente extendido en muchas regiones de Asia, África, Australia y América Latina, incluyendo Cuba. La cosecha del fruto del Nim se hace engorrosa debido a la altura que adquieren estos árboles a los pocos años de sembrados, siendo la cosecha mecánica mediante sacudidores de ramas o del tronco, una posible vía de solución práctica de esta problemática. En el presente trabajo se efectúa, mediante la simulación con el empleo de herramientas CAD-3D y programas de análisis por elementos finitos, un análisis modal abarcador de un número importante de modos de vibración del sistema fruto-pedúnculo del Nim. Como resultado del estudio se pudieron recomendar entornos de frecuencias de las vibraciones a aplicar con mejores expectativas durante la cosecha mecanizada de este producto.

  14. Astrid-2 EMMA Magnetic Calibration

    Merayo, José M.G.; Brauer, Peter; Risbo, Torben;


    magnetometer readings in each position were related to the field magnitudes from the Observatory, and a least squares fit for the 9 magnetometer calibration parameters was performed (3 offsets, 3 scale values and 3 inter-axes angles). After corrections for the magnetometer digital-to-analogue converters...... fit calibration parameters. Owing to time shortage, we did not evaluate the temperature coefficients of the flight sensor calibration parameters. However, this was done for an identical flight spare magnetometer sensor at the magnetic coil facility belonging to the Technical University of Braunschweig...

  15. Potencial del agua del suelo

    Bustamante Heliodoro


    La energía potencial del agua presenta diferencias de un punto del suelo a otro; esas diferencias son las que originan el movimiento del agua de acuerdo a la tendencia universal de la materia en el sentido de moverse de donde la energía potencial es mayor a donde dicha energía es menor. En el suelo el agua en consecuencia se mueve hacia donde su energía decrece hasta lograr su estado de equilibrio. Se desprende entonces que la cantidad de energía potencial absoluta contenida en el agua, no es...

  16. The calibration of radioprotection hardware

    After having recalled recent recommendations on dose limits on the basis of two radioprotection values (the equivalent and the efficient dose), this document indicates some characteristics of these values, and discusses how they are applied for individual monitoring and for area or ambient monitoring. It presents conventions aimed at simplifying radiation fields. Then, the author gives a precise overview of some general aspects concerning calibration operations: legal requirements, radioprotection hardware controls, calibration loop organisation (calibration definition, general physical values, reference radiation, conversion factors, and metrology), comparison between operational values and the protection value (irradiation geometries, conversion factors with respect to the geometries, comparison between efficient dose and operational values). He finally describes the calibration procedures: dosemeter location, energy response, angular response, flow rate response, uncertainties

  17. On chromatic and geometrical calibration

    Folm-Hansen, Jørgen


    The main subject of the present thesis is different methods for the geometrical and chromatic calibration of cameras in various environments. For the monochromatic issues of the calibration we present the acquisition of monochrome images, the classic monochrome aberrations and the various sources...... of non-uniformity of the illumination of the image plane. Only the image deforming aberrations and the non-uniformity of illumination are included in the calibration models. The topics of the pinhole camera model and the extension to the Direct Linear Transform (DLT) are described. It is shown how...... the correct interpolation method is described. For the chromatic issues of calibration we present the acquisition of colour and multi-spectral images, the chromatic aberrations and the various lens/camera based non-uniformities of the illumination of the image plane. It is described how the...

  18. Calibration of "Babyline" RP instruments


      If you have old RP instrumentation of the “Babyline” type, as shown in the photo, please contact the Radiation Protection Group (Joffrey Germa, 73171) to have the instrument checked and calibrated. Thank you. Radiation Protection Group

  19. Rotary mode system initial instrument calibration

    Johns, B.R.


    The attached report contains the vendor calibration procedures used for the initial instrument calibration of the rotary core sampling equipment. The procedures are from approved vendor information files.

  20. Rotary mode system initial instrument calibration

    The attached report contains the vendor calibration procedures used for the initial instrument calibration of the rotary core sampling equipment. The procedures are from approved vendor information files

  1. Beam Imaging and Luminosity Calibration

    AUTHOR|(CDS)2081126; Klute, Markus; Medlock, Catherine Aiko


    We discuss a method to reconstruct two-dimensional proton bunch densities using vertex distributions accumulated during LHC beam-beam scans. The x-y correlations in the beam shapes are studied and an alternative luminosity calibration technique is introduced. We demonstrate the method on simulated beam-beam scans and estimate the uncertainty on the luminosity calibration associated to the beam-shape reconstruction to be below 1%.

  2. CERI: Ionizing Radiation Calibration Centre

    The CERI has been granted by the National Bureau of Metrology (BNM) as an Ionizing Radiation Calibration Centre and as an Estimation and Qualification Centre for the ionizing radiation measurement devices. This article gives some information on the scope covered by the BNM's grant and on the various equipment on which the laboratory relies. It describes the calibration and estimation activities and mentions many kinds of services which are offered to the users mainly in the medical and industrial fields

  3. The Third VLBA Calibrator Survey

    Petrov, L.; Kovalev, Y. Y.; Fomalont, E.; Gordon, D


    This paper presents the third extension to the Very Large Baseline Array (VLBA) Calibrator Survey, containing 360 new sources not previously observed with very long baseline interferometry (VLBI). The survey, based on three 24 hour VLBA observing sessions, fills the areas on the sky above declination -45 degrees where the calibrator density is less than one source within a 4 degrees radius disk at any given direction. The positions were derived from astrometric analysis of the group delays de...

  4. Pressures Detector Calibration and Measurement

    Kumara, I Made Gita


    This is report of my first and second projects (of 3) in NA61. I did data taking and analysis in order to do calibration of pressure detectors and verified it. I analyzed the data by ROOT software using the C ++ programming language. The first part of my project was determination of calibration factor of pressure sensors. Based on that result, I examined the relation between pressure drop, gas flow rate of in paper filter and its diameter.

  5. WFC3: UVIS Dark Calibration

    Bourque, Matthew; Biretta, John A.; Anderson, Jay; Baggett, Sylvia M.; Gunning, Heather C.; MacKenty, John W.


    Wide Field Camera 3 (WFC3), a fourth-generation imaging instrument on board the Hubble Space Telescope (HST), has exhibited excellent performance since its installation during Servicing Mission 4 in May 2009. The UVIS detector, comprised of two e2v CCDs, is one of two channels available on WFC3 and is named for its ultraviolet and visible light sensitivity. We present the various procedures and results of the WFC3/UVIS dark calibration, which monitors the health and stability of the UVIS detector, provides characterization of hot pixels and dark current, and produces calibration files to be used as a correction for dark current in science images. We describe the long-term growth of hot pixels and the impacts that UVIS Charge Transfer Efficiency (CTE) losses, postflashing, and proximity to the readout amplifiers have on the population. We also discuss the evolution of the median dark current, which has been slowly increasing since the start of the mission and is currently ~6 e-/hr/pix, averaged across each chip. We outline the current algorithm for creating UVIS dark calibration files, which includes aggressive cosmic ray masking, image combination, and hot pixel flagging. Calibration products are available to the user community, typically 3-5 days after initial processing, through the Calibration Database System (CDBS). Finally, we discuss various improvements to the calibration and monitoring procedures. UVIS dark monitoring will continue throughout and beyond HST’s current proposal cycle.

  6. Calibration and Data Elaboration Procedure for Sky Irradiance Measurements

    Boi, Paolo; Tonna, Glauco; Dalu, Giuseppe; Nakajima, Teruyuki; Olivieri, Bruno; Pompei, Alberto; Campanelli, Monica; Rao, R.


    The problems encountered in the elaboration of measurements of direct and sky diffuse solar irradiance are the following: (1) to carry out the calibration for the direct irradiance, which consists in determining the direct irradiance at the upper limit of the atmosphere; (2) to carry out the calibration for the diffuse irradiance, which consists in determining the solid viewing angle of the sky radiometer; (3) to determine the input parameters, namely, ground albedo, real and imaginary parts of the aerosol refractive index, and aerosol radius range; and (4) to determine from the optical data the columnar aerosol optical depth and volume radius distribution. With experimental data and numerical simulations a procedure is shown that enables one to carry out the two calibrations needed for the sky radiometer, to determine a best estimate of the input parameters, and, finally, to obtain the average features of the atmospheric aerosols. An interesting finding is that inversion of only data of diffuse irradiance yields the same accuracy of result as data of both diffuse and direct irradiance; in this case, only calibration of the solid viewing angle of the sky radiometer is needed, thus shortening the elaboration procedure. Measurements were carried out in the Western Mediterranean Sea (Italy), in Tokyo (Japan), and in Ushuaia (Tierra del Fuego, Argentina); data were elaborated with a new software package, the Skyrad code, based on an efficient radiative transfer scheme.

  7. Evaluation of quality assurance calibration results based on repeated calibrations

    To ensure quality assurance of the calibration results, as indicated by the UNE-EN ISO / IEC 17025:2005 in paragraph 5.9, the laboratory has established procedures for quality control of its activity. Thus, the laboratory participates in both inter-laboratory intercomparison exercises, cycle through the entire range of radiation qualities reflected in the scope of its accreditation, such as intra-laboratory intercomparison exercises. In this case, repeat quarterly by two different operators both the calibration of an ionization chamber irradiation of a direct reading personal dosimeter.

  8. The KLOE Online Calibration System



    Based on all the features of the KLOE online software,the online calibration system performs current calibration quality checking in real time and starts automatically new calibration procedures when needed.Acalibration manager process controls the system,implementing the interface to the online system,receiving information from the run control and translating its state transitions to a separate state machine.It acts as a " calibration run controller"and performs failure recovery when requested by a set of process checkers.The core of the system is a multi-threaded OO histogram server that receives histogramming commands by remote processes and operates on local ROOT histograms.A client library and C,fortran and C++ application interface libraries allow the user to connect and define his own histogram or read histograms owned by others using an bool-like interface.Several calibration processes running in parallel in a destributed,multiplatform environment can fill the same histograms,allowing fast external information check.A monitor thread allow remote browsing for visual inspection,Pre-filtered data are read in nonprivileged spy mode from the data acquisition system via the Kloe Integrated Dataflow,privileged spy mode from the data acquisiton system via the Kole Integrated Dataflow.The main characteristics of the system are presented.

  9. Culturas del Mundo

    Benwell, Ann Fenger; Costa, Alberto; Waehle, Espen


    ’Culturas del mundo. Colecciones del Museo Nacional de Dinamarca’ with Ann Fenger Benwell in Culturas del Mundo. Colecciones del Museo Nacional de Dinamarca, ed. Silvia Sauquet, Fundación "la Caixa", Barcelona 2006, pp. 31-39......’Culturas del mundo. Colecciones del Museo Nacional de Dinamarca’ with Ann Fenger Benwell in Culturas del Mundo. Colecciones del Museo Nacional de Dinamarca, ed. Silvia Sauquet, Fundación "la Caixa", Barcelona 2006, pp. 31-39...

  10. The calibration system of the GERDA muon veto Cherenkov detector

    The GERDA experiment searches for neutrinoless double beta decay (0νββ). To achieve a sensitivity of 10-3counts/(keVkgy) or better within a specific region of interest (ROI), a good background identification is needed. Therefore GERDA is located in the LNGS (Laboratori Nationali del Gran Sasso) underground facility. In addition to the good rejection of cosmic muons due to the surrounding bedrocks, a dual muon veto system has to be used. For calibration and monitoring of the muon veto, two separate systems have been developed.

  11. Facility for dosimetric instrument calibration

    A structure is designed consisting of a rotary support of containers with radiation sources and of a rotary plug mounted above the sources. A support post with a slide rest and arms rotating around the post longitudinal axis is mounted in the centre of the container support. THe arms support instruments to be calibrated. The colimation cone of the respective source is directed to the tensor of the instrument being calibrated. The slide rest is balanced using a counterpoise mounted in the support post. Sources are not removed from the containers in source change during measurement. The support can hold more containers and the slide rest can support more instruments to be calibrated than the existing configurations. (M.D.). 2 figs

  12. Radiation protection dosimetry and calibrations

    At the SCK-CEN different specialised services are delivered for a whole range of external and internal customers in the radiation protection area. For the expertise group of radiation protection dosimetry and calibrations, these services are organized in four different laboratories: dosimetry, anthropogammametry, nuclear calibrations and non-nuclear calibrations. The services are given by a dedicated technical staff who has experience in the handling of routine and specialised cases. The scientific research that is performed by the expertise group makes sure that state-of-the-art techniques are being used, and that constant improvements and developments are implemented. Quality Assurance is an important aspect for the different services, and accreditation according national and international standards is achieved for all laboratories

  13. Optimal calibration of nuclear instrumentation

    Accurate knowledge of core power level is essential for the safe and efficient operation of nuclear power plants. Ionization chambers located outside the reactor core have the necessary reliability and response time characteristics and have been used extensively to indicate power level. The calibration of the ion chamber, and associated nuclear instrumentation (NI), has traditionally been based on the thermal power in the secondary coolant system. The usual NI calibration procedure consists of establishing steady-state operating conditions, calorimetrically determining the power at the secondary side of the steam generator, and adjusting the NI output to correspond to the measured thermal power. This study addresses certain questions including; (a) what sampling rate should be employed, (b) how many measurements are required, and (c) how can additional power level related information such as primary coolant loop measurements and knowledge of plant dynamics be included in the calibration procedure

  14. The Third VLBA Calibrator Survey

    Petrov, L A; Fomalont, E; Gordon, D


    This paper presents the third extension to the Very Large Baseline Array (VLBA) Calibrator Survey, containing 360 new sources not previously observed with very long baseline interferometry (VLBI). The survey, based on three 24 hour VLBA observing sessions, fills the areas on the sky above declination -45 deg where the calibrator density is less than one source within a 4 deg radius disk at any given direction. The positions were derived from astrometric analysis of the group delays determined at 2.3 and 8.4 GHz frequency bands using the Calc/Solve package. The VCS3 catalogue of source positions, plots of correlated flux density versus the length of projected baseline, contour plots and fits files of naturally weighted CLEAN images as well as calibrated visibility function files are available on the Web at

  15. Power calibrations for TRIGA reactors

    The purpose of this paper is to establish a framework for the calorimetric power calibration of TRIGA reactors so that reliable results can be obtained with a precision better than ± 5%. Careful application of the same procedures has produced power calibration results that have been reproducible to ± 1.5%. The procedures are equally applicable to the Mark I, Mark II and Mark III reactors as well as to reactors having much larger reactor tanks and to TRIGA reactors capable of forced cooling up to 3 MW in some cases and 15 MW in another case. In the case of forced cooled TRIGA reactors, the calorimetric power calibration is applicable in the natural convection mode for these reactors using exactly the same procedures as are discussed below for the smaller TRIGA reactors (< 2 MW)

  16. Performance standard for dose Calibrator

    Darmawati, S


    Dose calibrator is an instrument used in hospitals to determine the activity of radionuclide for nuclear medicine purposes. International Electrotechnical Commission (IEC) has published IEC 1303:1994 standard that can be used as guidance to test the performance of the instrument. This paper briefly describes content of the document,as well as explains the assessment that had been carried out to test the instrument accuracy in Indonesia through intercomparison measurement.Its is suggested that hospitals acquire a medical physicist to perform the test for its dose calibrator. The need for performance standard in the form of Indonesia Standard is also touched.

  17. Historia del experimento barométrico

    Rosa María Herrera


    Ciertos problemas hidráulicos planteados por el ingeniero Baliani hacía 1630 culminaron en el experimento barométrico realizado por el matemático Torricelli en 1644. Este último, ferviente galileano, pensaba la matemática como lenguaje con que la naturaleza habla a los hombres. Con este soporte teórico del mundo realizó un experimento físico con el que calibró la presión atmosférica y comprobó la existencia del vacío.

  18. Historia del experimento barométrico

    Rosa María Herrera


    Full Text Available Ciertos problemas hidráulicos planteados por el ingeniero Baliani hacía 1630 culminaron en el experimento barométrico realizado por el matemático Torricelli en 1644. Este último, ferviente galileano, pensaba la matemática como lenguaje con que la naturaleza habla a los hombres. Con este soporte teórico del mundo realizó un experimento físico con el que calibró la presión atmosférica y comprobó la existencia del vacío.

  19. Uniformity calibration for ICT image

    The uniformity of ICT image is impaired by beam hardening and the inconsistency of detector units responses. The beam hardening and the nonlinearity of the detector's output have been analyzed. The correction factors are determined experimentally by the detector's responses with different absorption length. The artifacts in the CT image of a symmetrical aluminium cylinder have been eliminated after calibration. (author)

  20. Measurement System and Calibration report

    Vesth, Allan; Kock, Carsten Weber

    The report describes power curve measurements carried out on a given wind turbine. The measurements are carried out in accordance to Ref. [1]. A site calibration has been carried out; see Ref. [2], and the measured flow correction factors for different wind directions are used in the present...

  1. Measurement System and Calibration report

    Gómez Arranz, Paula; Villanueva, Héctor

    This Measurement System & Calibration report is describing DTU’s measurement system installed at a specific wind turbine. A major part of the sensors has been installed by others (see [1]) the rest of the sensors have been installed by DTU. The results of the measurements, described in this report...

  2. Scalar Calibration of Vector Magnetometers

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz;


    The calibration parameters of a vector magnetometer are estimated only by the use of a scalar reference magnetometer. The method presented in this paper differs from those previously reported in its linearized parametrization. This allows the determination of three offsets or signals in the absence...

  3. Autonomous Attitude Sensor Calibration (ASCAL)

    Peterson, Chariya; Rowe, John; Mueller, Karl; Ziyad, Nigel


    In this paper, an approach to increase the degree of autonomy of flight software is proposed. We describe an enhancement of the Attitude Determination and Control System by augmenting it with self-calibration capability. Conventional attitude estimation and control algorithms are combined with higher level decision making and machine learning algorithms in order to deal with the uncertainty and complexity of the problem.

  4. Reliability-Based Code Calibration

    Faber, M.H.; Sørensen, John Dalsgaard


    The present paper addresses fundamental concepts of reliability based code calibration. First basic principles of structural reliability theory are introduced and it is shown how the results of FORM based reliability analysis may be related to partial safety factors and characteristic values...

  5. Recommended Inorganic Chemicals for Calibration.

    Moody, John R.; And Others


    All analytical techniques depend on the use of calibration chemicals to relate analyte concentration to instrumental parameters. Discusses the preparation of standard solutions and provides a critical evaluation of available materials. Lists elements by group and discusses the purity and uses of each. (MVL)

  6. Calibration of personnel dose meters

    Methods of calibrating both film and thermoluminescent dose meters (TLD) to photon and electron radiations are described. K fluorescent X-rays, heavily filtered X-ray beams, and isotope gamma rays are used at the Los Alamos calibration facility to measure the energy and angular response of radiation detectors over a photon energy range of 10 to 1000keV. Beam spectra, alignment, size and uniformity are discussed. The energy and angular response of dose meters to electrons is measured with beta-emitting isotopes varying in maximum energy from 770 to 2300keV. A free-air ionization chamber is the primary standard used in the measurement of photon radiation. Thimble-sized ionization chambers, calibrated to the free-air chamber, serve as secondary standards. Electron radiation is measured with an end-window ionization chamber having a 7mg/cm2 approximately tissue-equivalent plastic wall. Photon calibrations are performed with personnel dose meters in air, on a phantom, and in a phantom. If the personnel dose meter and secondary chamber are both in air, or both on or both in a phantom, the response of the LiF TLD chip, relative to the secondary chamber, is the same. However, the film dose meter shows a larger relative response on or in the phantom than in air. With beta sources, personnel dose meters are calibrated by exposing the dose meter either in air to a high-dose-rate 90Sr (90Y) source, or in contact with a low-dose-rate uranium source. The differences in personnel dose meter response observed between the two methods are discussed. The personnel dose meters are calibrated to determine penetrating doses by placing the secondary chamber 1cm deep in a phantom and the personnel dose meter on the surface, with a filter over the TLD to simulate 1cm depth. Non-penetrating dose calibrations are measured by placing both chamber and dose meter on the surface of the phantom. (author)

  7. Mathematical efficiency calibration in gamma spectroscopy

    Kaminski, S; Wilhelm, C


    Mathematical efficiency calibration with the LabSOCS software was introduced for two detectors in the measurement laboratory of the Central Safety Department of Forschungszentrum Karlsruhe. In the present contribution, conventional efficiency calibration of gamma spectroscopy systems and mathematical efficiency calibration with LabSOCS are compared with respect to their performance, uncertainties, expenses, and results. It is reported about the experience gained, and the advantages and disadvantages of both methods of efficiency calibration are listed. The results allow the conclusion to be drawn that mathematical efficiency calibration is a real alternative to conventional efficiency calibration of gamma spectroscopy systems as obtained by measurements of mixed gamma ray standard sources.

  8. Perfil del visitante del parque nacional Desierto del Carmen

    Elizabeth López Carré


    El artículo presenta los resultados obtenidos del proyecto de investigación titulado "Evaluación del potencial de comercialización y caracterización del visitante del Santo Desierto de Tenancingo", específicamente los correspondientes al perfil del visitante a dicho Parque Nacional. La metodología utilizada se sustenta en la investigación de mercados propuesta por Kinnear y Taylor. El resultado de dicha investigación permitió integrar el perfil del segmento de mercado religioso...

  9. Sources of uncertainty in vicarious calibration: understanding calibration target reflectance

    Anderson, K.; Milton, E. J.; Rollin, E.M.


    A field experiment investigated the hypothesis that the nadir reflectance of calibration surface substrates (asphalt and concrete) remains stable over a range of time-scales. Measurable differences in spectral reflectance factors were found over periods as short as 30 minutes. Multi-date reflectance measurements were compared using ANOVA and found to differ significantly (p = 0.001). Surface reflectance showed a relationship with the relative proportion of diffuse irradiance, over periods ...

  10. Extracting the MESA SR4000 calibrations

    Charleston, Sean A.; Dorrington, Adrian A.; Streeter, Lee; Cree, Michael J.


    Time-of-flight range imaging cameras are capable of acquiring depth images of a scene. Some algorithms require these cameras to be run in `raw mode', where any calibrations from the off-the-shelf manufacturers are lost. The calibration of the MESA SR4000 is herein investigated, with an attempt to reconstruct the full calibration. Possession of the factory calibration enables calibrated data to be acquired and manipulated even in "raw mode." This work is motivated by the problem of motion correction, in which the calibration must be separated into component parts to be applied at different stages in the algorithm. There are also other applications, in which multiple frequencies are required, such as multipath interference correction. The other frequencies can be calibrated in a similar way, using the factory calibration as a base. A novel technique for capturing the calibration data is described; a retro-reflector is used on a moving platform, which acts as a point source at a distance, resulting in planar waves on the sensor. A number of calibrations are retrieved from the camera, and are then modelled and compared to the factory calibration. When comparing the factory calibration to both the "raw mode" data, and the calibration described herein, a root mean squared error improvement of 51:3mm was seen, with a standard deviation improvement of 34:9mm.

  11. Radiometric Calibration of Osmi Imagery Using Solar Calibration

    Lee, Dong-Han; Kim, Yong-Seung


    OSMI (Ocean Scanning Multi-Spectral Imager) raw image data (Level 0) were acquired and radiometrically corrected. We have applied two methods, using solar & dark calibration data from OSMI sensor and comparing with the SeaWiFS data, to the radiometric correction of OSMI raw image data. First, we could get the values of the gain and the offset for each pixel and each band from comparing the solar & dark calibration data with the solar input radiance values, calculated from the transmittance, BRDF (Bidirectional Reflectance Distribution Function) and the solar incidence angle (¥â,¥è) of OSMI sensor. Applying this calibration data to OSMI raw image data, we got the two odd results, the lower value of the radiometric corrected image data than the expected value, and the Venetian Blind Effect in the radiometric corrected image data. Second, we could get the reasonable results from comparing OSMI raw image data with the SeaWiFS data, and get a new problem of OSMI sensor.

  12. Calibration and Validation of Measurement System

    Kofoed, Jens Peter; Riemann, Sven; Knapp, Wilfried

    The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype.......The report deals with the calibration of the measuring equipment on board the Wave Dragon, Nissum Bredning prototype....

  13. Attitude Sensor and Gyro Calibration for Messenger

    O'Shaughnessy, Daniel; Pittelkau, Mark E.


    The Redundant Inertial Measurement Unit Attitude Determination/Calibration (RADICAL(TM)) filter was used to estimate star tracker and gyro calibration parameters using MESSENGER telemetry data from three calibration events. We present an overview of the MESSENGER attitude sensors and their configuration is given, the calibration maneuvers are described, the results are compared with previous calibrations, and variations and trends in the estimated calibration parameters are examined. The warm restart and covariance bump features of the RADICAL(TM) filter were used to estimate calibration parameters from two disjoint telemetry streams. Results show that the calibration parameters converge faster with much less transient variation during convergence than when the filter is cold-started at the start of each telemetry stream.

  14. Automated Calibration of Dosimeters for Diagnostic Radiology

    Calibration of dosimeters for diagnostic radiology includes current and charge measurements, which are often repetitive. However, these measurements are usually done using modern electrometers, which are equipped with an RS-232 interface that enables instrument control from a computer. This paper presents an automated system aimed to the measurements for the calibration of dosimeters used in diagnostic radiology. A software application was developed, in order to achieve the acquisition of the electric charge readings, measured values of the monitor chamber, calculation of the calibration coefficient and issue of a calibration certificate. A primary data record file is filled and stored in the computer hard disk. The calibration method used was calibration by substitution. With this system, a better control over the calibration process is achieved and the need for human intervention is reduced. the automated system will be used in the calibration of dosimeters for diagnostic radiology at the Cuban Secondary Standard Dosimetry Laboratory of the Center for Radiation Protection and Hygiene. (Author)

  15. Automatic calibration system for pressure transducers


    Fifty-channel automatic pressure transducer calibration system increases quantity and accuracy for test evaluation calibration. The pressure transducers are installed in an environmental tests chamber and manifolded to connect them to a pressure balance which is uniform.

  16. Quality control for dose calibrators

    Nuclear medicine laboratories are required to assay samples of radioactivity to be administered to patients. Almost universally, these assays are accomplished by use of a well ionization chamber isotope calibrator. The Instituto de Radioprotecao e Dosimetria (Institute for Radiological Protection and Dosimetry) of the Comissao Nacional de Energia Nuclear (National Commission for Nuclear Energy) is carrying out a National Quality Control Programme in Nuclear Medicine, supported by the International Atomic Energy Agency. The assessment of the current needs and practices of quality control in the entire country of Brazil includes Dose Calibrators and Scintillation Cameras, but this manual is restricted to the former. Quality Control Procedures for these Instruments are described in this document together with specific recommendations and assessment of its accuracy. (author)

  17. Spinning angle optical calibration apparatus

    Beer, Stephen K.; Pratt, II, Harold R.


    An optical calibration apparatus is provided for calibrating and reproducing spinning angles in cross-polarization, nuclear magnetic resonance spectroscopy. An illuminated magnifying apparatus enables optical setting an accurate reproducing of spinning "magic angles" in cross-polarization, nuclear magnetic resonance spectroscopy experiments. A reference mark scribed on an edge of a spinning angle test sample holder is illuminated by a light source and viewed through a magnifying scope. When the "magic angle" of a sample material used as a standard is attained by varying the angular position of the sample holder, the coordinate position of the reference mark relative to a graduation or graduations on a reticle in the magnifying scope is noted. Thereafter, the spinning "magic angle" of a test material having similar nuclear properties to the standard is attained by returning the sample holder back to the originally noted coordinate position.

  18. Sky camera geometric calibration using solar observations

    Urquhart, B.; Kurtz, B; J. Kleissl


    A camera model and associated automated calibration procedure for stationary daytime sky imaging cameras is presented. The specific modeling and calibration needs are motivated by remotely deployed cameras used to forecast solar power production where cameras point skyward and use 180° fisheye lenses. Sun position in the sky and on the image plane provides a simple and automated approach to calibration; special equipment or calibration patterns are not required. Sun positio...

  19. Binary Classifier Calibration: Non-parametric approach

    Naeini, Mahdi Pakdaman; Cooper, Gregory F.; Hauskrecht, Milos


    Accurate calibration of probabilistic predictive models learned is critical for many practical prediction and decision-making tasks. There are two main categories of methods for building calibrated classifiers. One approach is to develop methods for learning probabilistic models that are well-calibrated, ab initio. The other approach is to use some post-processing methods for transforming the output of a classifier to be well calibrated, as for example histogram binning, Platt scaling, and is...

  20. WCDMA User Equipment Output Power Calibration

    Folkeson, Tea


    To save time in Flextronics high volume production, the time for test and calibration of mobile telephones need to be as short and accurate as possible. In the wideband code division multiple access (WCDMA) case, the output power calibration is the most critical calibration concerning accuracy. The aim with this thesis was to find a faster calibration method than the one that exists today and still retain accuracy. The Third Generation Partnership Project (3GPP) outlines the requirements of ...

  1. GREAT/SOFIA atmospheric calibration

    Guan, Xin; Stutzki, Jürgen; Graf, Urs U.; Güsten, Rolf; Okada, Yoko; Torres, Miguel Angel Requena; Simon, Robert; Wiesemeyer, Helmut


    The GREAT observations need frequency-selective calibration across the passband for the residual atmospheric opacity at flight altitude. At these altitudes the atmospheric opacity has both narrow and broad spectral features. To determine the atmospheric transmission at high spectral resolution, GREAT compares the observed atmospheric emission with atmospheric model predictions, and therefore depends on the validity of the atmospheric models. We discusse the problems identified in this compari...

  2. Automated Camera Array Fine Calibration

    Clouse, Daniel; Padgett, Curtis; Ansar, Adnan; Cheng, Yang


    Using aerial imagery, the JPL FineCalibration (JPL FineCal) software automatically tunes a set of existing CAHVOR camera models for an array of cameras. The software finds matching features in the overlap region between images from adjacent cameras, and uses these features to refine the camera models. It is not necessary to take special imagery of a known target and no surveying is required. JPL FineCal was developed for use with an aerial, persistent surveillance platform.

  3. Colour calibration for colour reproduction

    Emmel, P.; R. D. Hersch


    Due to the proliferation of low-cost colour devices (digital colour cameras, scanners, printers etc.) during the last few years, colour calibration has become an important issue. Such devices should faithfully reproduce colour images, but experience shows they don't. Among the main reasons, we note the diversity of acquisition, display and printing technologies which makes standardization difficult. Each device has a different gamut, i.e. a different set of colours that it can acquire or repr...

  4. Multidetector calibration for mass spectrometers

    The International Atomic Energy Agency's Safeguards Analytical Laboratory has performed calibration experiments to measure the different efficiencies among multi-Faraday detectors for a Finnigan-MAT 261 mass spectrometer. Two types of calibration experiments were performed: (1) peak-shift experiments and (2) peak-jump experiments. For peak-shift experiments, the ion intensities were measured for all isotopes of an element in different Faraday detectors. Repeated measurements were made by shifting the isotopes to various Faraday detectors. Two different peak-shifting schemes were used to measure plutonium (UK Pu5/92138) samples. For peak-jump experiments, ion intensities were measured in a reference Faraday detector for a single isotope and compared with those measured in the other Faraday detectors. Repeated measurements were made by switching back-and-forth between the reference Faraday detector and a selected Faraday detector. This switching procedure is repeated for all Faraday detectors. Peak-jump experiments were performed with replicate measurements of 239Pu, 187Re, and 238U. Detector efficiency factors were estimated for both peak-jump and peak-shift experiments using a flexible calibration model to statistically analyze both types of multidetector calibration experiments. Calculated detector efficiency factors were shown to depend on both the material analyzed and the experimental conditions. A single detector efficiency factor is not recommended for each detector that would be used to correct routine sample analyses. An alternative three-run peak-shift sample analysis should be considered. A statistical analysis of the data from this peak-shift experiment can adjust the isotopic ratio estimates for detector differences due to each sample analysis

  5. ASCAL: Autonomous Attitude Sensor Calibration

    Peterson, Chariya; Rowe, John; Mueller, Karl; Ziyad, Nigel


    Abstract In this paper, an approach to increase the degree of autonomy of flight software is proposed. We describe an enhancement of the Attitude Determination and Control System by augmenting it with self-calibration capability. Conventional attitude estimation and control algorithms are combined with higher level decision making and machine learning algorithms in order to deal with the uncertainty and complexity of the problem.

  6. Calibrating thermal behavior of electronics

    Chainer, Timothy J.; Parida, Pritish R.; Schultz, Mark D.


    A method includes determining a relationship between indirect thermal data for a processor and a measured temperature associated with the processor, during a calibration process, obtaining the indirect thermal data for the processor during actual operation of the processor, and determining an actual significant temperature associated with the processor during the actual operation using the indirect thermal data for the processor during actual operation of the processor and the relationship.

  7. PACS photometer calibration block analysis

    Moór, A; Kiss, Cs; Balog, Z; Billot, N; Marton, G


    The absolute stability of the PACS bolometer response over the entire mission lifetime without applying any corrections is about 0.5% (standard deviation) or about 8% peak-to-peak. This fantastic stability allows us to calibrate all scientific measurements by a fixed and time-independent response file, without using any information from the PACS internal calibration sources. However, the analysis of calibration block observations revealed clear correlations of the internal source signals with the evaporator temperature and a signal drift during the first half hour after the cooler recycling. These effects are small, but can be seen in repeated measurements of standard stars. From our analysis we established corrections for both effects which push the stability of the PACS bolometer response to about 0.2% (stdev) or 2% in the blue, 3% in the green and 5% in the red channel (peak-to-peak). After both corrections we still see a correlation of the signals with PACS FPU temperatures, possibly caused by parasitic h...

  8. Unassisted 3D camera calibration

    Atanassov, Kalin; Ramachandra, Vikas; Nash, James; Goma, Sergio R.


    With the rapid growth of 3D technology, 3D image capture has become a critical part of the 3D feature set on mobile phones. 3D image quality is affected by the scene geometry as well as on-the-device processing. An automatic 3D system usually assumes known camera poses accomplished by factory calibration using a special chart. In real life settings, pose parameters estimated by factory calibration can be negatively impacted by movements of the lens barrel due to shaking, focusing, or camera drop. If any of these factors displaces the optical axes of either or both cameras, vertical disparity might exceed the maximum tolerable margin and the 3D user may experience eye strain or headaches. To make 3D capture more practical, one needs to consider unassisted (on arbitrary scenes) calibration. In this paper, we propose an algorithm that relies on detection and matching of keypoints between left and right images. Frames containing erroneous matches, along with frames with insufficiently rich keypoint constellations, are detected and discarded. Roll, pitch yaw , and scale differences between left and right frames are then estimated. The algorithm performance is evaluated in terms of the remaining vertical disparity as compared to the maximum tolerable vertical disparity.

  9. Model Calibration in Watershed Hydrology

    Yilmaz, Koray K.; Vrugt, Jasper A.; Gupta, Hoshin V.; Sorooshian, Soroosh


    Hydrologic models use relatively simple mathematical equations to conceptualize and aggregate the complex, spatially distributed, and highly interrelated water, energy, and vegetation processes in a watershed. A consequence of process aggregation is that the model parameters often do not represent directly measurable entities and must, therefore, be estimated using measurements of the system inputs and outputs. During this process, known as model calibration, the parameters are adjusted so that the behavior of the model approximates, as closely and consistently as possible, the observed response of the hydrologic system over some historical period of time. This Chapter reviews the current state-of-the-art of model calibration in watershed hydrology with special emphasis on our own contributions in the last few decades. We discuss the historical background that has led to current perspectives, and review different approaches for manual and automatic single- and multi-objective parameter estimation. In particular, we highlight the recent developments in the calibration of distributed hydrologic models using parameter dimensionality reduction sampling, parameter regularization and parallel computing.

  10. Calibration of RB reactor power

    The first and only calibration of RB reactor power was done in 1962, and the obtained calibration ratio was used irrespective of the lattice pitch and core configuration. Since the RB reactor is being prepared for operation at higher power levels it was indispensable to reexamine the calibration ratio, estimate its dependence on the lattice pitch, critical level of heavy water and thickness of the side reflector. It was necessary to verify the reliability of control and dosimetry instruments, and establish neutron and gamma dose dependence on reactor power. Two series of experiments were done in June 1976. First series was devoted to tests of control and dosimetry instrumentation and measurements of radiation in the RB reactor building dependent on reactor power. Second series covered measurement of thermal and epithermal neuron fluxes in the reactor core and calculation of reactor power. Four different reactor cores were chosen for these experiments. Reactor pitches were 8, 8√2, and 16 cm with 40, 52 and 82 fuel channels containing 2% enriched fuel. Obtained results and analysis of these results are presented in this document with conclusions related to reactor safe operation

  11. A Numerical Simulator for Noise Calibration Studies

    Racette, Paul; Clune, Tom; Wong, Mark; Walker, David; Coakley, Kevin; Splett, Jolene; Rivers, Derick; Leonard, Robert; Boone, Ed


    Frequent calibration using noise references is used to reduce the effects of time varying fluctuations that naturally occur within sensitive radiometer receivers. Over the years, calibration architectures and processing algorithms have become more sophisticated. Predicting the performance of a given calibration architecture often requires expensive prototyping of hardware; optimizing processing algorithms for nonstationary fluctuations is a challenge. Measurement uncertainty is a figure of me...

  12. 48 CFR 908.7113 - Calibration services.


    ... 48 Federal Acquisition Regulations System 5 2010-10-01 2010-10-01 false Calibration services. 908... ACQUISITION PLANNING REQUIRED SOURCES OF SUPPLIES AND SERVICES Acquisition of Special Items 908.7113 Calibration services. Orders for calibration services may be placed with the National Institute of...

  13. MODIS Radiometric Calibration and Uncertainty Assessment

    Xiong, Xiaoxiong; Chiang, Vincent; Sun, Junqiang; Wu, Aisheng


    Since launch, Terra and Aqua MODIS have collected more than II and 9 years of datasets for comprehensive studies of the Earth's land, ocean, and atmospheric properties. MODIS observations are made in 36 spectral bands: 20 reflective solar bands (RSB) and 16 thermal emissive bands (TEB). Compared to its heritage sensors, MODIS was developed with very stringent calibration and uncertainty requirements. As a result, MODIS was designed and built with a set of state of the art on-board calibrators (OBC), which allow key sensor performance parameters and on-orbit calibration coefficients to be monitored and updated if necessary. In terms of its calibration traceability, MODIS RSB calibration is reflectance based using an on-board solar diffuser (SD) and the TEB calibration is radiance based using an on-board blackbody (BB). In addition to on-orbit calibration coefficients derived from its OBC, calibration parameters determined from sensor pre-launch calibration and characterization are used in both the RSB and TEB calibration and retrieval algorithms. This paper provides a brief description of MODIS calibration methodologies and discusses details of its on-orbit calibration uncertainties. It assesses uncertainty contributions from individual components and differences between Terra and Aqua MODIS due to their design characteristics and on-orbit periormance. Also discussed in this paper is the use of MODIS LIB uncertainty index CUI) product.

  14. New method for calibration of sun photometers

    H.H.Asadov; I.G.Chobanzadeh


    A new method for calibration of sun photometers based on Bouguer-Beer law is proposed.The developed basic equation of calibration makes it possible to formulate the derivative methods of calibration on the basis of photometric measurements upon optical air masses,the ratio of which is an integer number.

  15. Net analyte signal calculation for multivariate calibration

    Ferre, J.; Faber, N.M.


    A unifying framework for calibration and prediction in multivariate calibration is shown based on the concept of the net analyte signal (NAS). From this perspective, the calibration step can be regarded as the calculation of a net sensitivity vector, whose length is the amount of net signal when the

  16. Calibration of nuclear medicine gamma counters

    In this paper the practical problem of nuclear medicine gamma counters calibration has been solved by using dose calibrators CRC-15R with standard error ±5%. The samples from technetium generators have been measured both by dose calibrators CRC-15R and gamma counter ICN Gamma 3.33 taking into account decay correction. Only the linear part of the curve has practical meaning. The advantage of this procedure satisfies the requirements from international standards: the calibration of sources used for medical exposure be traceable to a standard dosimetry laboratory and radiopharmaceuticals for nuclear medicine procedures be calibrated in terms of activity of the radiopharmaceutical to be administered. (author)

  17. Borexino calibrations: Hardware, Methods, and Results

    Back, H; Benziger, J; Bick, D; Bonfini, G; Bravo, D; Avanzini, M Buizza; Caccianiga, B; Cadonati, L; Calaprice, F; Carraro, C; Cavalcante, P; Chavarria, A; D'Angelo, D; Davini, S; Derbin, A; Etenko, A; von Feilitzsch, F; Fernandes, G; Fomenko, K; Franco, D; Galbiati, C; Gazzana, S; Ghiano, C; Giammarchi, M; Goeger-Neff, M; Goretti, A; Grandi, L; Guardincerri, E; Hardy, S; Ianni, Aldo; Ianni, Andrea; Kayunov, A; Kidner, S; Kobychev, V; Korablev, D; Korga, G; Koshio, Y; Kryn, D; Laubenstein, M; Lewke, T; Litvinovich, E; Loer, B; Lombardi, F; Lombardi, P; Ludhova, L; Machulin, I; Manecki, S; Maneschg, W; Manuzio, G; Meindl, Q; Meroni, E; Miramonti, L; Misiaszek, M; Montanari, D; Mosteiro, P; Muratova, V; Oberauer, L; Obolensky, M; Ortica, F; Otis, K; Pallavicini, M; Papp, L; Perasso, L; Perasso, S; Pocar, A; Raghavan, R S; Ranucci, G; Razeto, A; Re, A; Romani, A; Rossi, N; Rountree, D; Sabelnikov, A; Saldanha, R; Salvo, C; Schonert, S; Simgen, H; Skorokhvatov, M; Smirnov, O; Sotnikov, A; Sukhotin, S; Suvorov, Y; Tartaglia, R; Testera, G; Vignaud, D; Vogelaar, R B; Winter, J; Wojcik, M; Wright, A; Wurm, M; Xu, J; Zaimidoroga, O; Zavatarelli, S; Zuzel, G


    Borexino was the first experiment to detect solar neutrinos in real-time in the sub-MeV region. In order to achieve high precision in the determination of neutrino rates, the detector design includes an internal and an external calibration system. This paper describes both calibration systems and the calibration campaigns that were carried out in the period between 2008 and 2011. We discuss some of the results and show that the calibration procedures preserved the radiopurity of the scintillator. The calibrations provided a detailed understanding of the detector response and led to a significant reduction of the systematic uncertainties in the Borexino measurements.

  18. Calibration of detector sensitivity in positron cameras

    An improved method for calibrating detector sensitivities in a positron camera has been developed. The calibration phantom is a cylinder of activity placed near the center of the camera and fully within the field of view. The calibration data is processed in such a manner that the following two important properties are achieved. The estimate of a detector sensitivity is unaffected by the sensitivities of the other detectors. The estimates are insensitive to displacements of the calibrating phantom from the camera center. Both of these properties produce a more accurate detector calibration

  19. ATLAS calorimeters energy calibration for jets

    The calibration of ATLAS barrel calorimeters (including pre shower system, electromagnetic Liquid Argon calorimeter and scintillating hadron tile calorimeter) was done by standard calibration and weighting technique approaches. The standard calibration gives the bad linearity for hadron non compensated calorimeter. The calibration with weighting technique, in comparison with standard calibration, restores linearity and improves energy resolution up to (σ/E)2 = (38.6%/√E)2 + (1.5%)2 for η 0.6. 6 refs., 4 figs., 1 tab

  20. Muon Energy Calibration of the MINOS Detectors

    Miyagawa, Paul S.


    MINOS is a long-baseline neutrino oscillation experiment designed to search for conclusive evidence of neutrino oscillations and to measure the oscillation parameters precisely. MINOS comprises two iron tracking calorimeters located at Fermilab and Soudan. The Calibration Detector at CERN is a third MINOS detector used as part of the detector response calibration programme. A correct energy calibration between these detectors is crucial for the accurate measurement of oscillation parameters. This thesis presents a calibration developed to produce a uniform response within a detector using cosmic muons. Reconstruction of tracks in cosmic ray data is discussed. This data is utilized to calculate calibration constants for each readout channel of the Calibration Detector. These constants have an average statistical error of 1.8%. The consistency of the constants is demonstrated both within a single run and between runs separated by a few days. Results are presented from applying the calibration to test beam particles measured by the Calibration Detector. The responses are calibrated to within 1.8% systematic error. The potential impact of the calibration on the measurement of oscillation parameters by MINOS is also investigated. Applying the calibration reduces the errors in the measured parameters by {approx} 10%, which is equivalent to increasing the amount of data by 20%.

  1. MODIS Instrument Operation and Calibration Improvements

    Xiong, X.; Angal, A.; Madhavan, S.; Link, D.; Geng, X.; Wenny, B.; Wu, A.; Chen, H.; Salomonson, V.


    Terra and Aqua MODIS have successfully operated for over 14 and 12 years since their respective launches in 1999 and 2002. The MODIS on-orbit calibration is performed using a set of on-board calibrators, which include a solar diffuser for calibrating the reflective solar bands (RSB) and a blackbody for the thermal emissive bands (TEB). On-orbit changes in the sensor responses as well as key performance parameters are monitored using the measurements of these on-board calibrators. This paper provides an overview of MODIS on-orbit operation and calibration activities, and instrument long-term performance. It presents a brief summary of the calibration enhancements made in the latest MODIS data collection 6 (C6). Future improvements in the MODIS calibration and their potential applications to the S-NPP VIIRS are also discussed.

  2. Calibration of the Cherenkov Telescope Array

    Gaug, Markus; Berge, David; Reyes, Raquel de los; Doro, Michele; Foerster, Andreas; Maccarone, Maria Concetta; Parsons, Dan; van Eldik, Christopher


    The construction of the Cherenkov Telescope Array is expected to start soon. We will present the baseline methods and their extensions currently foreseen to calibrate the observatory. These are bound to achieve the strong requirements on allowed systematic uncertainties for the reconstructed gamma-ray energy and flux scales, as well as on the pointing resolution, and on the overall duty cycle of the observatory. Onsite calibration activities are designed to include a robust and efficient calibration of the telescope cameras, and various methods and instruments to achieve calibration of the overall optical throughput of each telescope, leading to both inter-telescope calibration and an absolute calibration of the entire observatory. One important aspect of the onsite calibration is a correct understanding of the atmosphere above the telescopes, which constitutes the calorimeter of this detection technique. It is planned to be constantly monitored with state-of-the-art instruments to obtain a full molecular and...

  3. Multi-Axis Accelerometer Calibration System

    Finley, Tom; Parker, Peter


    A low-cost, portable, and simplified system has been developed that is suitable for in-situ calibration and/or evaluation of multi-axis inertial measurement instruments. This system overcomes facility restrictions and maintains or improves the calibration quality for users of accelerometer-based instruments with applications in avionics, experimental wind tunnel research, and force balance calibration applications. The apparatus quickly and easily positions a multi-axis accelerometer system into a precisely known orientation suitable for in-situ quality checks and calibration. In addition, the system incorporates powerful and sophisticated statistical methods, known as response surface methodology and statistical quality control. These methods improve calibration quality, reduce calibration time, and allow for increased calibration frequency, which enables the monitoring of instrument stability over time.

  4. Laser Calibration of an Impact Disdrometer

    Lane, John E.; Kasparis, Takis; Metzger, Philip T.; Jones, W. Linwood


    A practical approach to developing an operational low-cost disdrometer hinges on implementing an effective in situ adaptive calibration strategy. This calibration strategy lowers the cost of the device and provides a method to guarantee continued automatic calibration. In previous work, a collocated tipping bucket rain gauge was utilized to provide a calibration signal to the disdrometer's digital signal processing software. Rainfall rate is proportional to the 11/3 moment of the drop size distribution (a 7/2 moment can also be assumed, depending on the choice of terminal velocity relationship). In the previous case, the disdrometer calibration was characterized and weighted to the 11/3 moment of the drop size distribution (DSD). Optical extinction by rainfall is proportional to the 2nd moment of the DSD. Using visible laser light as a means to focus and generate an auxiliary calibration signal, the adaptive calibration processing is significantly improved.

  5. The Palomar Testbed Interferometer Calibrator Catalog

    Van Belle, G T; Creech-Eakman, M J; Coyne, J; Boden, A F; Akeson, R L; Ciardi, D R; Rykoski, K M; Thompson, R R; Lane, B F


    The Palomar Testbed Interferometer (PTI) archive of observations between 1998 and 2005 is examined for objects appropriate for calibration of optical long-baseline interferometer observations - stars that are predictably point-like and single. Approximately 1,400 nights of data on 1,800 objects were examined for this investigation. We compare those observations to an intensively studied object that is a suitable calibrator, HD217014, and statistically compare each candidate calibrator to that object by computing both a Mahalanobis distance and a Principal Component Analysis. Our hypothesis is that the frequency distribution of visibility data associated with calibrator stars differs from non-calibrator stars such as binary stars. Spectroscopic binaries resolved by PTI, objects known to be unsuitable for calibrator use, are similarly tested to establish detection limits of this approach. From this investigation, we find more than 350 observed stars suitable for use as calibrators (with an additional $\\approx 1...

  6. Pulse-based internal calibration of polarimetric SAR

    Dall, Jørgen; Skou, Niels; Christensen, Erik Lintz

    Internal calibration greatly diminishes the dependence on calibration target deployment compared to external calibration. Therefore the Electromagnetics Institute (EMI) at the Technical University of Denmark (TUD) has equipped its polarimetric SAR, EMISAR, with several calibration loops and devel...

  7. Protocols for calibrating multibeam sonar

    Foote, Kenneth G.; Chu, Dezhang; Hammar, Terence R.; Baldwin, Kenneth C.; Mayer, Larry A.; Hufnagle, Lawrence C. jr.; Jech, J. Michael


    Development of protocols for calibrating multibeam sonar by means of the standard-target method is documented. Particular systems used in the development work included three that provide the water-column signals, namely the SIMRAD SM2000/90- and 200-kHz sonars and RESON SeaBat 8101 sonar, with operating frequency of 240 kHz. Two facilities were instrumented specifically for the work: a sea well at the Woods Hole Oceanographic Institution and a large, indoor freshwater tank at the University o...

  8. Calibration of TOB+ Thermometer's Cards

    Banitt, Daniel


    Motivation - Under the new upgrade of the CMS detector the working temperature of the trackers had been reduced to -27 Celsius degrees. Though the thermal sensors themselves (Murata and Fenwal thermistors) are effective at these temperatures, the max1542 PLC (programmable logic controller) cards, interpreting the resistance of the thermal sensors into DC counts usable by the DCS (detector control system), are not designed for these temperatures in which the counts exceed their saturation and therefor had to be replaced. In my project I was in charge of handling the emplacement and calibration of the new PLC cards to the TOB (tracker outer barrel) control system.

  9. Scintillation camera brightness calibration apparatus

    Circuitry is described for calibrating the brightness of a cathode ray tube display and recording apparatus comprising: 1) intensity control means for adjusting the intensity of the cathode ray tube beam; 2) light sensitive means disposed to receive light emitted from the cathode ray tube and generating a first electrical signal having a magnitude dependent upon the intensity of the emitted light; 3) reference signal generating means for generating a second electrical signal of predetermined magnitude; and 4) electrical signal comparison means coupled to the light sensitive means and the reference signal generating means for comparing the magnitude of the first and second electrical signals. (author)

  10. Scalar Calibration of Vector Magnetometers

    Merayo, José M.G.; Brauer, Peter; Primdahl, Fritz; Petersen, Jan Raagaard; Nielsen, Otto V


    uniquely the parameters for a given data set. Therefore, a magnetometer may be characterized inexpensively in the Earth's magnetic-field environment. This procedure has been used successfully in the pre-flight calibration of the state-of-the-art magnetometers on board the magnetic mapping satellites Orsted...... of a magnetic field, three scale factors for normalization of the axes and three non-orthogonality angles which build up an orthogonal system intrinsically in the sensor. The advantage of this method compared with others lies in its linear least squares estimator, which finds independently and...

  11. Percutaneous transhepatic fine calibre cholangioscopy

    Klose, K.J.; Thelen, M.; Schild, H.H.


    Percutaneous transhepatic fine calibre cholangioscopy is described. A specially developed transparent instrument is used, which makes it unnecessary to have a steerable endoscope; visualisation of the biliary tree is of diagnostic value for assessing biliary stenoses due to tumours (choice of biopsy site, extent of intraductal radiation therapy) and for the assessment of anastomoses between the biliary system and the gut (condition of the mucosa, stenoses). In the presence of percutaneous biliary drainage, the method has little utility. Extension of the method for treating intraductal stones may be possible if in future a useful lumen can be added to the fine endoscope.

  12. del cambio

    Alejandro Canales


    Full Text Available El artículo analiza las propuestas y los compromisos en materia de cobertura y de financiamiento en el periodo 2000-2006. En particular, revisa las modificaciones y los ajustes de los tres principales documentos en los que se expresaron las iniciativas: el de la Asociación Nacional de Universidades e Instituciones de Educación Superior, el del equipo de la transición y el programa sectorial. Los resultados muestran que las cifras de atención a la demanda educativa e inversión quedaron lejos de las metas autoimpuestas por el gobierno federal y que será necesario realizar mayores esfuerzos para atender tales desafíos.

  13. MODIS Solar Reflective Calibration Traceability

    Xiong, Xiaoxiong; Butler, Jim


    Long-term climate data records often consist of observations made by multiple sensors. It is, therefore, extremely important to have instrument overlap, to be able to track instrument stability, to quantify, measurement uncertainties, and to establish absolute scale traceable to the International System of Units (SI). The Moderate Resolution Imaging Spectroradiometer (MODIS) is a key instrument for both the Terra and Aqua missions, which were launched in December 1999 and May 2002, respectively. It has 20 reflective solar bands (RSB) with wavelengths from 0.41 to 2.2 micrometers and observes the Earth at three nadir spatial resolutions: 0.25km, 0.5km, and 1km. MODIS RSB on-orbit calibration is reflectance based with reference to the bidirectional reflectance factor (BRF) of its on-board solar diffuser (SD). The SD BRF characterization was made pre-launch by the instrument vendor using reference samples traceable directly to the National Institute of Standards and Technology (NIST). On-orbit SD reflectance degradation is tracked by an on-board solar diffuser monitor (SDSM). This paper provides details of this calibration chain, from prelaunch to on-orbit operation, and associated uncertainty assessments. Using MODIS as an example, this paper also discusses challenges and key design requirements for future missions developed for accurate climate studies.

  14. MODIS solar reflective calibration traceability

    Xiong, Xiaoxiong; Butler, Jim


    Long-term climate data records often consist of observations made by multiple sensors. It is, therefore, extremely important to have instrument overlap, to be able to track instrument stability, to quantify measurement uncertainties, and to establish an absolute measurement scale traceable to the International System of Units (SI). The Moderate Resolution Imaging Spectroradiometer (MODIS) is a key instrument for both the Terra and Aqua missions, which were launched in December 1999 and May 2002, respectively. It has 20 reflective solar bands (RSB) with wavelengths from 0.41 to 2.2μm and observes the Earth at three nadir spatial resolutions: 0.25km, 0.5km, and 1km. MODIS RSB on-orbit calibration is reflectance based with reference to the bi-directional reflectance factor (BRF) of its on-board solar diffuser (SD). The SD BRF characterization was made pre-launch by the instrument vendor using reference samples traceable directly to the National Institute of Standards and Technology (NIST). On-orbit SD reflectance degradation is tracked by an on-board solar diffuser stability monitor (SDSM). This paper provides details of this calibration chain, from pre-launch to on-orbit operation, and associated uncertainty assessments. Using MODIS as an example, this paper also discusses challenges and key design requirements for future missions developed for accurate climate studies.

  15. Another look at volume self-calibration: calibration and self-calibration within a pinhole model of Scheimpflug cameras

    Cornic, Philippe; Illoul, Cédric; Cheminet, Adam; Le Besnerais, Guy; Champagnat, Frédéric; Le Sant, Yves; Leclaire, Benjamin


    We address calibration and self-calibration of tomographic PIV experiments within a pinhole model of cameras. A complete and explicit pinhole model of a camera equipped with a 2-tilt angles Scheimpflug adapter is presented. It is then used in a calibration procedure based on a freely moving calibration plate. While the resulting calibrations are accurate enough for Tomo-PIV, we confirm, through a simple experiment, that they are not stable in time, and illustrate how the pinhole framework can be used to provide a quantitative evaluation of geometrical drifts in the setup. We propose an original self-calibration method based on global optimization of the extrinsic parameters of the pinhole model. These methods are successfully applied to the tomographic PIV of an air jet experiment. An unexpected by-product of our work is to show that volume self-calibration induces a change in the world frame coordinates. Provided the calibration drift is small, as generally observed in PIV, the bias on the estimated velocity field is negligible but the absolute location cannot be accurately recovered using standard calibration data.

  16. Herschel SPIRE FTS Relative Spectral Response Calibration

    Fulton, Trevor; Baluteau, Jean-Paul; Benielli, Dominique; Imhof, Peter; Lim, Tanya; Lu, Nanyao; Marchili, Nicola; Naylor, David; Polehampton, Edward; Swinyard, Bruce; Valtchanov, Ivan


    Herschel/SPIRE Fourier transform spectrometer (FTS) observations contain emission from both the Herschel Telescope and the SPIRE Instrument itself, both of which are typically orders of magnitude greater than the emission from the astronomical source, and must be removed in order to recover the source spectrum. The effects of the Herschel Telescope and the SPIRE Instrument are removed during data reduction using relative spectral response calibration curves and emission models. We present the evolution of the methods used to derive the relative spectral response calibration curves for the SPIRE FTS. The relationship between the calibration curves and the ultimate sensitivity of calibrated SPIRE FTS data is discussed and the results from the derivation methods are compared. These comparisons show that the latest derivation methods result in calibration curves that impart a factor of between 2 and 100 less noise to the overall error budget, which results in calibrated spectra for individual observations whose n...

  17. Absolute calibration in vivo measurement systems

    Lawrence Livermore National Laboratory (LLNL) is currently investigating a new method for obtaining absolute calibration factors for radiation measurement systems used to measure internally deposited radionuclides in vivo. Absolute calibration of in vivo measurement systems will eliminate the need to generate a series of human surrogate structures (i.e., phantoms) for calibrating in vivo measurement systems. The absolute calibration of in vivo measurement systems utilizes magnetic resonance imaging (MRI) to define physiological structure, size, and composition. The MRI image provides a digitized representation of the physiological structure, which allows for any mathematical distribution of radionuclides within the body. Using Monte Carlo transport codes, the emission spectrum from the body is predicted. The in vivo measurement equipment is calibrated using the Monte Carlo code and adjusting for the intrinsic properties of the detection system. The calibration factors are verified using measurements of existing phantoms and previously obtained measurements of human volunteers. 8 refs

  18. Calibration Monitor for Dark Energy Experiments

    Kaiser, M. E.


    The goal of this program was to design, build, test, and characterize a flight qualified calibration source and monitor for a Dark Energy related experiment: ACCESS - 'Absolute Color Calibration Experiment for Standard Stars'. This calibration source, the On-board Calibration Monitor (OCM), is a key component of our ACCESS spectrophotometric calibration program. The OCM will be flown as part of the ACCESS sub-orbital rocket payload in addition to monitoring instrument sensitivity on the ground. The objective of the OCM is to minimize systematic errors associated with any potential changes in the ACCESS instrument sensitivity. Importantly, the OCM will be used to monitor instrument sensitivity immediately after astronomical observations while the instrument payload is parachuting to the ground. Through monitoring, we can detect, track, characterize, and thus correct for any changes in instrument senstivity over the proposed 5-year duration of the assembled and calibrated instrument.

  19. Research of Camera Calibration Based on DSP

    Zheng Zhang


    Full Text Available To take advantage of the high-efficiency and stability of DSP in the data processing and the functions of OpenCV library, this study brought forward a scheme that camera calibration in DSP embedded system calibration. An arithmetic of camera calibration based on OpenCV is designed by analyzing the camera model and lens distortion. The transplantation of EMCV to DSP is completed and the arithmetic of camera calibration is migrated and optimized based on the CCS development environment and the DSP/BIOS system. On the premise of realizing calibration function, this arithmetic improves the efficiency of program execution and the precision of calibration and lays the foundation for further research of the visual location based on DSP embedded system.

  20. Calibration of beta dosimeter and personal dosimeter

    This paper introduces ISO standard ISO 6980 which prepared especially for the calibration of beta dosimeter and personal dosimeter. The standard has three aspects including method of production of reference beta particle radiations, calibration fundamentals related to basic quantities characterizing the radiation field, and calibration of area and personal dosemters and the determination of their response as a function of beta radiation energy and angle of incidence. Here particular emphasis is placed on the determination of basic quantity of tissue absorbed dose at a depth of 0.07 mm in the tissue slab phantom and calibration procedure by mean of the calibration quantity of directional dose equivalent H'(0.07, Ω) and personal dose equivalent Hp (0.07, Ω). Finally, combined standard uncertainty for the determination of absorbed dose rate and component uncertainties of calibration is given as examples. (authors)

  1. Calibration biases in logical reasoning tasks

    Guillermo Macbeth; Alfredo López Alonso; Eugenia Razumiejczyk; Rodrigo Sosa; Carolina Pereyra; Humberto Fernández


    The aim of this contribution is to present an experimental study about calibration in deductive reasoning tasks. Calibration is defi ned as the empirical convergence or divergence between the objective and the subjective success. The underconfi dence bias is understood as the dominance of the former over the latter. The hypothesis of this study states that the form of the propositions presented in the experiment is critical for calibration phenomena. Affi rmative and negative propositions are...

  2. Calibration of Nacelle-based Lidar instrument

    Yordanova, Ginka; Courtney, Michael

    This report presents the result of the lidar calibration performed for a two-beam nacelle based lidar at DTU’s test site for large wind turbines at Høvsøre, Denmark. Calibration is here understood as the establishment of a relation between the reference wind speed measurements with measurement...... uncertainties provided by measurement standard and corresponding lidar wind speed indications with associated measurement uncertainties. The lidar calibration concerns the 10 minute mean wind speed measurements....

  3. On Single-scanline Camera Calibration

    Horaud, Radu; Mohr, Roger; Lorecki, Boguslaw


    A method for calibrating single scanline CCD cameras is described. It is shown that the more classical 2D camera calibration techniques are necessary but not sufficient for solving the 1D camera calibration problem. A model for single scanline cameras is proposed, and a two-step procedure for estimating its parameters is provided. It is also shown how the extrinsic camera parameters can be determined geometrically without making explicit the intrinsic camera parameters. The accuracy of the ca...

  4. Camera calibration from surfaces of revolution

    Wong, KYK; Mendonça, PRS; Cipolla, R.


    This paper addresses the problem of calibrating a pinhole camera from images of a surface of revolution. Camera calibration is the process of determining the intrinsic or internal parameters (i.e., aspect ratio, focal length, and principal point) of a camera, and it is important for both motion estimation and metric reconstruction of 3D models. In this paper, a novel and simple calibration technique is introduced, which is based on exploiting the symmetry of images of surfaces of revolution. ...

  5. Calibration metrology for fixed irradiation sensors

    After having recalled the regulatory and technical framework of the calibration of radioprotection measurement instruments, and outlined some technical and operational constraints, the authors report the development of an in situ calibration methodology, i.e. without displacement of the sensor. After the presentation of the calibration chain (from the measurement given by a fixed sensor to the reference value given by a primary standard), they indicate the definition and calculation of the different calibration coefficients allowing the linking up of the different levels, and also the taking of uncertainties into account. They finally report the validation of results

  6. The calibration system for the GERDA experiment

    The GERDA experiment uses the neutrinoless double beta decay to probe three fundamental questions in neutrino physics - Are they Dirac or Majorana particles? What is their absolute mass? What is the mass hierarchy of the three generations? In my talk I present the calibration system for the Ge semiconductor diodes enriched in Ge-76. The system is used to set the energy scale and calibrate the pulse shapes which will be used to further reject background events. The lowest possible background is crucial for the whole experiment and therefore the calibration system must not interfere with the data acquisition phase while at the same time operate efficiently during the calibration runs.

  7. Sloan Digital Sky Survey Photometric Calibration Revisited

    Marriner, John; /Fermilab


    The Sloan Digital Sky Survey calibration is revisited to obtain the most accurate photometric calibration. A small but significant error is found in the flat-fielding of the Photometric telescope used for calibration. Two SDSS star catalogs are compared and the average difference in magnitude as a function of right ascension and declination exhibits small systematic errors in relative calibration. The photometric transformation from the SDSS Photometric Telescope to the 2.5 m telescope is recomputed and compared to synthetic magnitudes computed from measured filter bandpasses.

  8. Sloan Digital Sky Survey Photometric Calibration Revisited

    Marriner, J.


    The Sloan Digital Sky Survey calibration is revisited to obtain the most accurate photometric calibration. A small but significant error is found in the flat-fielding of the Photometric telescope used for calibration. Two SDSS star catalogs are compared and the average difference in magnitude as a function of right ascension and declination exhibits small systematic errors in relative calibration. The photometric transformation from the SDSS Photometric Telescope to the 2.5 m telescope is recomputed and compared to synthetic magnitudes computed from measured filter bandpasses.

  9. The long term stability of lidar calibrations

    Courtney, Michael; Gayle Nygaard, Nicolai

    Wind lidars are now used extensively for wind resource measurements. One of the requirements for the data to be accepted in support of project financing (so-called ‘banka-bility’) is to demonstrate the long-term stability of lidar cali-brations. Calibration results for six Leosphere WindCube li-dars......-ters pertaining in the different calibration periods. This is supported by sliding-window analyses of one lidar at one location where the same order of variation is observed as between pre-service and post-service calibrations....

  10. Research of Camera Calibration Based on DSP

    Zheng Zhang; Yukun Wan; Lixin Cai


    To take advantage of the high-efficiency and stability of DSP in the data processing and the functions of OpenCV library, this study brought forward a scheme that camera calibration in DSP embedded system calibration. An arithmetic of camera calibration based on OpenCV is designed by analyzing the camera model and lens distortion. The transplantation of EMCV to DSP is completed and the arithmetic of camera calibration is migrated and optimized based on the CCS development environment and the ...