WorldWideScience

Sample records for breast x-ray ct

  1. Toward optimal X-ray flux utilization in breast CT

    DEFF Research Database (Denmark)

    Jørgensen, Jakob Heide; Hansen, Per Christian; Sidky, Emil Y.

    2011-01-01

    A realistic computer-simulation of a breast computed tomography (CT) system and subject is constructed. The model is used to investigate the optimal number of views for the scan given a fixed total X-ray fluence. The reconstruction algorithm is based on accurate solution to a constrained......, TVminimization problem, which has received much interest recently for sparse-view CT data....

  2. Attenuation correction of myocardial SPECT images with X-ray CT. Effects of registration errors between X-ray CT and SPECT

    International Nuclear Information System (INIS)

    Takahashi, Yasuyuki; Murase, Kenya; Mochizuki, Teruhito; Motomura, Nobutoku

    2002-01-01

    Attenuation correction with an X-ray CT image is a new method to correct attenuation on SPECT imaging, but the effect of the registration errors between CT and SPECT images is unclear. In this study, we investigated the effects of the registration errors on myocardial SPECT, analyzing data from a phantom and a human volunteer. Registerion (fusion) of the X-ray CT and SPECT images was done with standard packaged software in three dimensional fashion, by using linked transaxial, coronal and sagittal images. In the phantom study, and X-ray CT image was shifted 1 to 3 pixels on the x, y and z axes, and rotated 6 degrees clockwise. Attenuation correction maps generated from each misaligned X-ray CT image were used to reconstruct misaligned SPECT images of the phantom filled with 201 Tl. In a human volunteer, X-ray CT was acquired in different conditions (during inspiration vs. expiration). CT values were transferred to an attenuation constant by using straight lines; an attenuation constant of 0/cm in the air (CT value=-1,000 HU) and that of 0.150/cm in water (CT value=0 HU). For comparison, attenuation correction with transmission CT (TCT) data and an external γ-ray source ( 99m Tc) was also applied to reconstruct SPECT images. Simulated breast attenuation with a breast attachment, and inferior wall attenuation were properly corrected by means of the attenuation correction map generated from X-ray CT. As pixel shift increased, deviation of the SPECT images increased in misaligned images in the phantom study. In the human study, SPECT images were affected by the scan conditions of the X-ray CT. Attenuation correction of myocardial SPECT with an X-ray CT image is a simple and potentially beneficial method for clinical use, but accurate registration of the X-ray CT to SPECT image is essential for satisfactory attenuation correction. (author)

  3. X-ray examination for breast cancer: Benefit versus risk

    International Nuclear Information System (INIS)

    Dalrymple, G.V.; Baker, M.L.

    1984-01-01

    Cancer of the breast is the most common malignancy afflicting American women. According to the American Cancer Society, one of 11 women (9 percent) born in the United States today, will develop breast cancer in her lifetime. Twenty-seven percent of all cancers in women and 19 percent of all cancer deaths in women are attributable to breast cancer. In 1982, 112,000 women were found to have cancer of the breast, and 37,000 women died from breast cancer. X-ray examinations of the breast are of considerable value in the diagnosis of breast cancer. This may be especially true in the asymptomatic patient who does not have a palpable mass. These x-ray examinations, however, are associated with both a finite though small risk of induction of cancer of the breasts and even smaller risk of death from cancer of the breast. This chapter presents a brief review of cancer of the breast and discusses the value of diagnostic studies, including x-ray mammography; the benefits and risks associated with x-ray examinations; and the future potential of computed tomography (CT) and ultrasound as imaging modalities in the detection of breast cancer

  4. Phase-contrast X-ray CT

    Energy Technology Data Exchange (ETDEWEB)

    Momose, Atsushi [Hitachi Ltd., Saitama (Japan). Advanced Research Laboratory; Takeda, Tohoru; Itai, Yuji

    1995-12-01

    Phase-contrast X-ray computed tomography (CT) enabling the observation of biological soft tissues without contrast enhancement has been developed. The X-ray phase shift caused by an object is measured and input to a standard CT reconstruction algorithm. A thousand times increase in the image sensitivity to soft tissues is achieved compared with the conventional CT using absorption contrast. This is because the X-ray phase shift cross section of light elements is about a thousand times larger than the absorption cross section. The phase shift is detected using an X-ray interferometer and computer analyses of interference patterns. Experiments were performed using a synchrotron X-ray source. Excellent image sensitivity is demonstrated in the observation of cancerous rabbit liver. The CT images distinguish cancer lesion from normal liver tissue and, moreover, visualize the pathological condition in the lesion. Although the X-ray energy employed and the present observation area size are not suitable for medical applications as they are, phase-contrast X-ray CT is promising for investigating the internal structure of soft tissue which is almost transparent for X-rays. The high sensitivity also provides the advantage of reducing X-ray doses. (author).

  5. Material Discriminated X-Ray CT System by Using New X-Ray Imager with Energy Discriminate Function

    Directory of Open Access Journals (Sweden)

    Toru Aoki

    2008-04-01

    Full Text Available Material discriminated X-ray CT system has been constructed by using conventional X-ray tube (white X-ray source and photon-counting X-ray imager as an application with energy band detection. We have already reported material identify X-ray CT using K-shell edge method elsewhere. In this report the principle of material discrimination was adapted the separation of electron-density and atomic number from attenuation coefficient mapping in X-ray CT reconstructed image in two wavelength X-ray CT method using white X-ray source and energy discriminated X-ray imager by using two monochrome X-ray source method. The measurement phantom was prepared as four kinds material rods (Carbon(C, Iron(Fe, Copper(Cu, Titanium(Ti rods of 3mm-diameter inside an aluminum(Al rod of 20mm-diameter. We could observed material discriminated X-ray CT reconstructed image, however, the discrimination properties were not good than two monochrome X-ray CT method. This results was could be explained because X-ray scattering, beam-hardening and so on based on white X-ray source, which could not observe in two monochrome X-ray CT method. However, since our developed CdTe imager can be detect five energy-bands at the same time, we can use multi-band analysis to decrease the least square error margin. We will be able to obtain more high separation in atomic number mapping in X-ray CT reconstructed image by using this system.

  6. Quantitative breast tissue characterization using grating-based x-ray phase-contrast imaging

    Science.gov (United States)

    Willner, M.; Herzen, J.; Grandl, S.; Auweter, S.; Mayr, D.; Hipp, A.; Chabior, M.; Sarapata, A.; Achterhold, K.; Zanette, I.; Weitkamp, T.; Sztrókay, A.; Hellerhoff, K.; Reiser, M.; Pfeiffer, F.

    2014-04-01

    X-ray phase-contrast imaging has received growing interest in recent years due to its high capability in visualizing soft tissue. Breast imaging became the focus of particular attention as it is considered the most promising candidate for a first clinical application of this contrast modality. In this study, we investigate quantitative breast tissue characterization using grating-based phase-contrast computed tomography (CT) at conventional polychromatic x-ray sources. Different breast specimens have been scanned at a laboratory phase-contrast imaging setup and were correlated to histopathology. Ascertained tumor types include phylloides tumor, fibroadenoma and infiltrating lobular carcinoma. Identified tissue types comprising adipose, fibroglandular and tumor tissue have been analyzed in terms of phase-contrast Hounsfield units and are compared to high-quality, high-resolution data obtained with monochromatic synchrotron radiation, as well as calculated values based on tabulated tissue properties. The results give a good impression of the method’s prospects and limitations for potential tumor detection and the associated demands on such a phase-contrast breast CT system. Furthermore, the evaluated quantitative tissue values serve as a reference for simulations and the design of dedicated phantoms for phase-contrast mammography.

  7. A survey on coordinate metrology using dimensional X-ray CT

    International Nuclear Information System (INIS)

    Matsuzaki, Kazuya

    2016-01-01

    X-ray computed tomography (X-ray CT) has been occupying indispensable position in geometrical and dimensional measurements in industry, which is capable of measuring both external and internal dimensions of industrial products. Since dimensional X-ray CT has problems about ensuring traceability and estimating uncertainty, requirement of developing measurement standard for dimensional X-ray CT is increasing. Some of national metrology institutes (NMIs) including NMIJ have been working on developing measurement standard. In this report, the background of coordinate metrology using dimensional X-ray CT is reviewed. Then, measurement error sources are discussed. Finally, the plan to develop high accuracy dimensional X-ray CT is presented. (author)

  8. Three-dimensional monochromatic x-ray CT

    Science.gov (United States)

    Saito, Tsuneo; Kudo, Hiroyuki; Takeda, Tohoru; Itai, Yuji; Tokumori, Kenji; Toyofuku, Fukai; Hyodo, Kazuyuki; Ando, Masami; Nishimura, Ktsuyuki; Uyama, Chikao

    1995-08-01

    In this paper, we describe a 3D computed tomography (3D CT) using monochromatic x-rays generated by synchrotron radiation, which performs a direct reconstruction of 3D volume image of an object from its cone-beam projections. For the develpment of 3D CT, scanning orbit of x-ray source to obtain complete 3D information about an object and corresponding 3D image reconstruction algorithm are considered. Computer simulation studies demonstrate the validities of proposed scanning method and reconstruction algorithm. A prototype experimental system of 3D CT was constructed. Basic phantom examinations and specific material CT image by energy subtraction obtained in this experimental system are shown.

  9. Performance evaluation of X-ray CT using visible scintillation light

    International Nuclear Information System (INIS)

    Kodama, Kiyoyuki; Hamada, Minoru; Suzuki, Tamotsu; Hashimoto, Masatoshi; Hanada, Takashi; Ide, Tatsuya; Maruyama, Koichi

    2004-01-01

    We proposed a new method of performance evaluation for X-ray CT using visible scintillation light and examined its usefulness in this study. When we scanned a plastic scintillator disk in a gantry opening of the X-ray CT, we could observe visible scintillation light. The rotation of the light-emitting area of the disk corresponded to that of the X-ray tube. We were able to record the scintillation light by digital video camera. By analyzing the area of visible scintillation light, the rotation speed of the X-ray tube, angular spread of the X-ray beam, uniformity of the incident X-rays, and change in X-ray energy were measured. No other method is available to obtain the above parameters of X-ray CT during a single CT scan. In the measurements of the uniformity of incident X-rays and change of X-ray energy, our method showed good accuracy in detecting the attenuation caused by the couch between the X-ray tube and the plastic scintillator disc. The proposed method is inexpensive and easy-to-use. We conclude that the method is a useful tool for performance evaluation as well as a maintenance tool for X-ray CT. (author)

  10. An X-Ray computed tomography/positron emission tomography system designed specifically for breast imaging.

    Science.gov (United States)

    Boone, John M; Yang, Kai; Burkett, George W; Packard, Nathan J; Huang, Shih-ying; Bowen, Spencer; Badawi, Ramsey D; Lindfors, Karen K

    2010-02-01

    Mammography has served the population of women who are at-risk for breast cancer well over the past 30 years. While mammography has undergone a number of changes as digital detector technology has advanced, other modalities such as computed tomography have experienced technological sophistication over this same time frame as well. The advent of large field of view flat panel detector systems enable the development of breast CT and several other niche CT applications, which rely on cone beam geometry. The breast, it turns out, is well suited to cone beam CT imaging because the lack of bones reduces artifacts, and the natural tapering of the breast anteriorly reduces the x-ray path lengths through the breast at large cone angle, reducing cone beam artifacts as well. We are in the process of designing a third prototype system which will enable the use of breast CT for image guided interventional procedures. This system will have several copies fabricated so that several breast CT scanners can be used in a multi-institutional clinical trial to better understand the role that this technology can bring to breast imaging.

  11. X-ray and CT findings of costal eosinophilic granuloma

    International Nuclear Information System (INIS)

    Tu Zhanhai; Lin Zhengyu; Chen Yiguang

    2010-01-01

    Objective: To study the X-ray and CT features of costal eosinophilic granuloma for a better understanding. Methods: Eight patients with costal eosinophilic granuloma proved by surgery or biopsy were analyzed retrospectively. All patients had X-ray plain film, 6 patients had CT examination, including a case of enhanced CT scan. Results: All 8 lesions were solitary. Six lesions were in the anterior rib and 2 in the posterior rib. On X-ray, all case showed single cavity and oval lesion with clear boundary. On CT images, 5 lesions demonstrated expansile destruction of bone with cortical bone thinning, and 3 were osteolystic destruction with soft tissue mass around. On the patient with enhanced CT scan, the lesions showed a moderate and uniform enhancement. Conclusion: The X-ray and CT findings of costal eosinophilic granuloma are characteristic. (authors)

  12. Cranial x-ray CT and MRI in congenital muscular dystrophy

    International Nuclear Information System (INIS)

    Horikawa, Hirosei; Konishi, Toshihiko; Konagaya, Masaaki; Mano, Yukio; Takayanagi, Tetsuya

    1988-01-01

    The involvements of central nervous system in those cases of congenital muscular dystrophy (CMD), especially in Fukuyama type CMD, have been observed both radiologically and pathologically. The recent development of MRI made it easier to detect fine structural changes in brain matter than the X-ray CT. Then, we tried to evaluate the central nervous system abnormalities of six cases of CMD by both X-ray CT and MRI. In one case, X-ray CT revealed diffuse hypodensity of cerebral white matter, and MRI showed high intensity on long spin-echo image and low intensity on inversion-recovery image. In another case, X-ray CT showed no abnormal findings, but long spin-echo image revealed two high intensity spots in cerebral white matter. In other four cases, brain atrophy was demonstrated by X-ray CT and/or MRI, one case of these patients had bilateral congenital arachnoid cysts in the middle cranial fossa and hypogenesis of temporal lobes. Although we could not demonstrate polymicrogyria and agyria in all cases by MRI, white matter changes and structural changes were revealed more clearly than X-ray CT. The combination of X-ray CT and MRI seems to make a noteworthy contribution to estimate the central nervous system abnormalities in CMD. (author)

  13. TU-EF-207-03: Advances in Stationary Breast Tomosynthesis Using Distributed X-Ray Sources

    Energy Technology Data Exchange (ETDEWEB)

    Zhou, O. [The University of North Carolina at Chapel Hill (United States)

    2015-06-15

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation.

  14. TU-EF-207-03: Advances in Stationary Breast Tomosynthesis Using Distributed X-Ray Sources

    International Nuclear Information System (INIS)

    Zhou, O.

    2015-01-01

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation

  15. Development of high resolution x-ray CT technique for irradiated fuel assembly

    Energy Technology Data Exchange (ETDEWEB)

    Ishimi, Akihiro; Katsuyama, Kozo; Maeda, Koji; Asaga, Takeo [Japan Atomic Energy Agency, Oarai Research and Development Center, Oarai, Ibaraki (Japan)

    2012-03-15

    High X-ray CT technique was developed to observe the irradiation performance of FBR fuel assembly and MOX fuel. In this technique, the high energy X-ray pulse (12MeV) was used synchronizing detection system with the X-ray pulse to reduce the effect of the gamma ray emissions from the irradiated fuel assembly. In this study, this technique was upgraded to obtain high resolution X-ray CT image. In this upgrading, the collimator which had slit width of 0.1 mm and X-ray detector of a highly sensitive silicon semiconductor detector (100 channels) was introduced in the X-ray CT system. As a result of these developments, high resolution X-ray CT images could be obtained on the transverse cross section of irradiated fuel assembly. (author)

  16. Low-Dose Contrast-Enhanced Breast CT Using Spectral Shaping Filters: An Experimental Study.

    Science.gov (United States)

    Makeev, Andrey; Glick, Stephen J

    2017-12-01

    Iodinated contrast-enhanced X-ray imaging of the breast has been studied with various modalities, including full-field digital mammography (FFDM), digital breast tomosynthesis (DBT), and dedicated breast CT. Contrast imaging with breast CT has a number of advantages over FFDM and DBT, including the lack of breast compression, and generation of fully isotropic 3-D reconstructions. Nonetheless, for breast CT to be considered as a viable tool for routine clinical use, it would be desirable to reduce radiation dose. One approach for dose reduction in breast CT is spectral shaping using X-ray filters. In this paper, two high atomic number filter materials are studied, namely, gadolinium (Gd) and erbium (Er), and compared with Al and Cu filters currently used in breast CT systems. Task-based performance is assessed by imaging a cylindrical poly(methyl methacrylate) phantom with iodine inserts on a benchtop breast CT system that emulates clinical breast CT. To evaluate detectability, a channelized hoteling observer (CHO) is used with sums of Laguerre-Gauss channels. It was observed that spectral shaping using Er and Gd filters substantially increased the dose efficiency (defined as signal-to-noise ratio of the CHO divided by mean glandular dose) as compared with kilovolt peak and filter settings used in commercial and prototype breast CT systems. These experimental phantom study results are encouraging for reducing dose of breast CT, however, further evaluation involving patients is needed.

  17. Dual resolution cone beam breast CT: A feasibility study

    International Nuclear Information System (INIS)

    Chen Lingyun; Shen Youtao; Lai, Chao-Jen; Han Tao; Zhong Yuncheng; Ge Shuaiping; Liu Xinming; Wang Tianpeng; Yang, Wei T.; Whitman, Gary J.; Shaw, Chris C.

    2009-01-01

    Purpose: In this study, the authors investigated the feasibility of a dual resolution volume-of-interest (VOI) cone beam breast CT technique and compared two implementation approaches in terms of dose saving and scatter reduction. Methods: With this technique, a lead VOI mask with an opening is inserted between the x-ray source and the breast to deliver x-ray exposure to the VOI while blocking x rays outside the VOI. A CCD detector is used to collect the high resolution projection data of the VOI. Low resolution cone beam CT (CBCT) images of the entire breast, acquired with a flat panel (FP) detector, were used to calculate the projection data outside the VOI with the ray-tracing reprojection method. The Feldkamp-Davis-Kress filtered backprojection algorithm was used to reconstruct the dual resolution 3D images. Breast phantoms with 180 μm and smaller microcalcifications (MCs) were imaged with both FP and FP-CCD dual resolution CBCT systems, respectively. Two approaches of implementing the dual resolution technique, breast-centered approach and VOI-centered approach, were investigated and evaluated for dose saving and scatter reduction with Monte Carlo simulation using a GEANT4 package. Results: The results showed that the breast-centered approach saved more breast absorbed dose than did VOI-centered approach with similar scatter reduction. The MCs in fatty breast phantom, which were invisible with FP CBCT scan, became visible with the FP-CCD dual resolution CBCT scan. Conclusions: These results indicate potential improvement of the image quality inside the VOI with reduced breast dose both inside and outside the VOI.

  18. Hard X-ray Microscopic Imaging Of Human Breast Tissues

    Science.gov (United States)

    Park, Sung H.; Kim, Hong T.; Kim, Jong K.; Jheon, Sang H.; Youn, Hwa S.

    2007-01-01

    X-ray microscopy with synchrotron radiation will be a useful tool for innovation of x-ray imaging in clinical and laboratory settings. It helps us observe detailed internal structure of material samples non-invasively in air. And, it also has the potential to solve some tough problems of conventional breast imaging if it could evaluate various conditions of breast tissue effectively. A new hard x-ray microscope with a spatial resolution better than 100 nm was installed at Pohang Light Source, a third generation synchrotron radiation facility in Pohang, Korea. The x-ray energy was set at 6.95 keV, and the x-ray beam was monochromatized by W/B4C monochromator. Condenser and objective zone plates were used as x-ray lenses. Zernike phase plate next to condenser zone plate was introduced for improved contrast imaging. The image of a sample was magnified 30 times by objective zone plate and 20 times by microscope objective, respectively. After additional 10 times digital magnification, the total magnifying power was up to 6000 times in the end. Phase contrast synchrotron images of 10-μm-thick female breast tissue of the normal, fibroadenoma, fibrocystic change and carcinoma cases were obtained. By phase contrast imaging, hard x-rays enable us to observe many structures of breast tissue without sample preparations such as staining or fixation.

  19. Hard X-ray Microscopic Imaging Of Human Breast Tissues

    International Nuclear Information System (INIS)

    Park, Sung H.; Kim, Hong T.; Kim, Jong K.; Jheon, Sang H.; Youn, Hwa S.

    2007-01-01

    X-ray microscopy with synchrotron radiation will be a useful tool for innovation of x-ray imaging in clinical and laboratory settings. It helps us observe detailed internal structure of material samples non-invasively in air. And, it also has the potential to solve some tough problems of conventional breast imaging if it could evaluate various conditions of breast tissue effectively. A new hard x-ray microscope with a spatial resolution better than 100 nm was installed at Pohang Light Source, a third generation synchrotron radiation facility in Pohang, Korea. The x-ray energy was set at 6.95 keV, and the x-ray beam was monochromatized by W/B4C monochromator. Condenser and objective zone plates were used as x-ray lenses. Zernike phase plate next to condenser zone plate was introduced for improved contrast imaging. The image of a sample was magnified 30 times by objective zone plate and 20 times by microscope objective, respectively. After additional 10 times digital magnification, the total magnifying power was up to 6000 times in the end. Phase contrast synchrotron images of 10-μm-thick female breast tissue of the normal, fibroadenoma, fibrocystic change and carcinoma cases were obtained. By phase contrast imaging, hard x-rays enable us to observe many structures of breast tissue without sample preparations such as staining or fixation

  20. A framework for breast cancer visualization using augmented reality x-ray vision technique in mobile technology

    Science.gov (United States)

    Rahman, Hameedur; Arshad, Haslina; Mahmud, Rozi; Mahayuddin, Zainal Rasyid

    2017-10-01

    Breast Cancer patients who require breast biopsy has increased over the past years. Augmented Reality guided core biopsy of breast has become the method of choice for researchers. However, this cancer visualization has limitations to the extent of superimposing the 3D imaging data only. In this paper, we are introducing an Augmented Reality visualization framework that enables breast cancer biopsy image guidance by using X-Ray vision technique on a mobile display. This framework consists of 4 phases where it initially acquires the image from CT/MRI and process the medical images into 3D slices, secondly it will purify these 3D grayscale slices into 3D breast tumor model using 3D modeling reconstruction technique. Further, in visualization processing this virtual 3D breast tumor model has been enhanced using X-ray vision technique to see through the skin of the phantom and the final composition of it is displayed on handheld device to optimize the accuracy of the visualization in six degree of freedom. The framework is perceived as an improved visualization experience because the Augmented Reality x-ray vision allowed direct understanding of the breast tumor beyond the visible surface and direct guidance towards accurate biopsy targets.

  1. Is there any advantage of CT based 3-dimensional conformal planning over conventional orthogonal x-ray based planning in HDR brachytherapy in breast cancer

    International Nuclear Information System (INIS)

    Biswal, B.M.; Idris, N.R.; Zakaria, A.B.; Khairul, N.

    2003-01-01

    The conventional brachytherapy dose calculation is based on a particular brachytherapy rule or individual dosimetry based on the reconstruction of the sources from the orthogonal films. In the recent years many centers are using CT based 3D conformal brachytherapy in order to improve the dosimetric outcome of a given plan. Here we would like to present our experience on the use of both techniques to deliver HDR interstitial brachytherapy as boost in early breast cancer. From January 2001 to January 2003, we treated 4 breast cancer patients using conventional orthogonal x-rays and CT scan in 3 cases for the treatment plan. All patients received an external beam radiotherapy dose of 46 Gy in 23 fractions over 4.5 weeks to the whole breast using 6 MV photon beam. Subsequently the primary lesion was supplimented with HDR brachytherapy to a dose of 2.5 Gy BID for 3 consecutive days using a (192)Ir microSelectronHDR. The dose prescription was individualized to encompass the tumor volume with a 10 mm margin. The differences of the dosimetric outcome were compared. All patients completed above schedule of radiotherapy. The primary was implanted with single plane in 3 patients and multiplane implant in 4 patients. Orthogonal x-ray based localization was performed in 4 patients and CT scan based localization in 3 cases. Three patients were implanted single plane and 4 patients with multiplane implants with a median catheter number of 9 (range 6-14). The 3D conformal dose optimization was performed using Nucletron planning system (Plato). The mean 100% and 150% isodose volume was 67.3 cm 3 and 31.25cm 3 respectively. The identification of primary tumor volume, organ at risk, and identification of afterloading catheters were superior in CT based plan than conventional planning. CT scan based 3D conformal brachytherapy planning give better identification of tumor volume and its curvature, decrease the time to identify the sources and evaluate the radiation dose to organs at

  2. Reconstructing misaligned x-ray CT data

    Energy Technology Data Exchange (ETDEWEB)

    Divin, C. J. [Lawrence Livermore National Lab. (LLNL), Livermore, CA (United States)

    2016-10-24

    Misalignment errors for x-ray computed tomography (CT) systems can manifest as artifacts and a loss of spatial and contrast resolution. To mitigate artifacts, significant effort is taken to determine the system geometry and minimizing any residual error in the system alignment. This project improved our ability to post-correct data which was acquired on a misaligned CT system.

  3. Introduction of X-ray CT application in geotechnical engineering - theory and practice

    International Nuclear Information System (INIS)

    Otani, J; Watanabe, Y; Chevalier, B

    2010-01-01

    In this paper, fundamental issues on the use of X-ray CT to geomaterials were discussed, which were the contents of what you have to know about the use of X-ray CT, especially for geomaterials or granular materials. Then, two of our recent studies were introduced, which are 1) the use of X-ray CT data to quantitative measurements such as deformation and strains and 2) comparison of X-ray CT results with numerical analysis. Here, the triaxial compression test on sand was conducted and the image analysis technique called 'Digital Image Correlation' was used to measure displacement and strain fields in the soils using a large number of CT images. For the second topic, one of the advantages on the use of X-ray CT is to investigate the behavior in three dimensions. Here, so called trap door test was conducted with the materials of glass beads and the shape of the square door. DEM analysis was also conducted for the simulation of this behaviour and both results were compared. Based on all the results in this paper, the effectiveness of the use of X-ray CT in geomechanics and geotechnical engineering was revealed.

  4. X-ray scatter correction method for dedicated breast computed tomography: improvements and initial patient testing

    International Nuclear Information System (INIS)

    Ramamurthy, Senthil; D’Orsi, Carl J; Sechopoulos, Ioannis

    2016-01-01

    A previously proposed x-ray scatter correction method for dedicated breast computed tomography was further developed and implemented so as to allow for initial patient testing. The method involves the acquisition of a complete second set of breast CT projections covering 360° with a perforated tungsten plate in the path of the x-ray beam. To make patient testing feasible, a wirelessly controlled electronic positioner for the tungsten plate was designed and added to a breast CT system. Other improvements to the algorithm were implemented, including automated exclusion of non-valid primary estimate points and the use of a different approximation method to estimate the full scatter signal. To evaluate the effectiveness of the algorithm, evaluation of the resulting image quality was performed with a breast phantom and with nine patient images. The improvements in the algorithm resulted in the avoidance of introduction of artifacts, especially at the object borders, which was an issue in the previous implementation in some cases. Both contrast, in terms of signal difference and signal difference-to-noise ratio were improved with the proposed method, as opposed to with the correction algorithm incorporated in the system, which does not recover contrast. Patient image evaluation also showed enhanced contrast, better cupping correction, and more consistent voxel values for the different tissues. The algorithm also reduces artifacts present in reconstructions of non-regularly shaped breasts. With the implemented hardware and software improvements, the proposed method can be reliably used during patient breast CT imaging, resulting in improvement of image quality, no introduction of artifacts, and in some cases reduction of artifacts already present. The impact of the algorithm on actual clinical performance for detection, diagnosis and other clinical tasks in breast imaging remains to be evaluated. (paper)

  5. Dedicated breast CT: radiation dose for circle-plus-line trajectory

    International Nuclear Information System (INIS)

    Vedantham, Srinivasan; Shi, Linxi; Karellas, Andrew; Noo, Frederic

    2012-01-01

    Purpose: Dedicated breast CT prototypes used in clinical investigations utilize single circular source trajectory and cone-beam geometry with flat-panel detectors that do not satisfy data-sufficiency conditions and could lead to cone beam artifacts. Hence, this work investigated the glandular dose characteristics of a circle-plus-line trajectory that fulfills data-sufficiency conditions for image reconstruction in dedicated breast CT. Methods: Monte Carlo-based computer simulations were performed using the GEANT4 toolkit and was validated with previously reported normalized glandular dose coefficients for one prototype breast CT system. Upon validation, Monte Carlo simulations were performed to determine the normalized glandular dose coefficients as a function of x-ray source position along the line scan. The source-to-axis of rotation distance and the source-to-detector distance were maintained constant at 65 and 100 cm, respectively, in all simulations. The ratio of the normalized glandular dose coefficient at each source position along the line scan to that for the circular scan, defined as relative normalized glandular dose coefficient (RD g N), was studied by varying the diameter of the breast at the chest wall, chest-wall to nipple distance, skin thickness, x-ray beam energy, and glandular fraction of the breast. Results: The RD g N metric when stated as a function of source position along the line scan, relative to the maximum length of line scan needed for data sufficiency, was found to be minimally dependent on breast diameter, chest-wall to nipple distance, skin thickness, glandular fraction, and x-ray photon energy. This observation facilitates easy estimation of the average glandular dose of the line scan. Polynomial fit equations for computing the RD g N and hence the average glandular dose are provided. Conclusions: For a breast CT system that acquires 300-500 projections over 2π for the circular scan, the addition of a line trajectory with equal

  6. The upper limits of the SNR in radiography and CT with polyenergetic x-rays

    International Nuclear Information System (INIS)

    Shikhaliev, Polad M

    2010-01-01

    The aim of the study is to determine the upper limits of the signal-to-noise ratio (SNR) in radiography and computed tomography (CT) with polyenergetic x-ray sources. In x-ray imaging, monoenergetic x-rays provide a higher SNR compared to polyenergetic x-rays. However, the SNR in polyenergetic x-ray imaging can be increased when a photon-counting detector is used and x-rays are optimally weighted according to their energies. For a particular contrast/background combination and at a fixed x-ray entrance skin exposure, the SNR in energy-weighting x-ray imaging depends on tube voltage and can be maximized by selecting the optimal tube voltage. The SNR in energy-weighted x-ray images acquired at this optimal tube voltage is the highest SNR that can be achieved with polyenergetic x-ray sources. The optimal tube voltages and the highest SNR were calculated and compared to the SNR of monoenergetic x-ray imaging. Monoenergetic, energy-weighting polyenergetic and energy-integrating polyenergetic x-ray imagings were simulated at a fixed entrance skin exposure of 20 mR. The tube voltages varied in the range of 30-140 kVp with 10 kV steps. Contrast elements of CaCO 3 , iodine, adipose and tumor with thicknesses of 280 mg cm -2 , 15 mg cm -2 , 1 g cm -2 and 1 g cm -2 , respectively, inserted in a soft tissue background with 10 cm and 20 cm thicknesses, were used. The energy weighting also improves the contrast-to-noise ratio (CNR) in CT when monoenergetic CT projections are optimally weighted prior to CT reconstruction (projection-based weighting). Alternatively, monoenergetic CT images are reconstructed, optimally weighted and composed to yield a final CT image (image-based weighting). Both projection-based and image-based weighting methods improve the CNR in CT. An analytical approach was used to determine which of these two weighting methods provides the upper limit of the CNR in CT. The energy-weighting method was generalized and expanded as a weighting method applicable in

  7. X-ray scatter signatures for enhanced breast imaging

    Energy Technology Data Exchange (ETDEWEB)

    Kidane, Ghirmay; Speller, Robert; Royle, Gary [Medical Physics and Bioengineering Department, University College Landon, 11-20 Capper Street, London WC1E 6JA (United Kingdom)

    1999-12-31

    Conventional mammographic imaging suffers from a low specificity. The main cause is the small difference in the x-ray attenuation properties of healthy and diseased tissue leading to poor contrast in the image. It has been observed that additional information on breast tissue type can be obtained from x-ray diffraction effects. A study of excised normal and neoplastic breast tissue samples using x-ray diffraction apparatus has been observed that significant differences exist in the measured spectra between carcinoma and healthy tissue adjacent to the carcinoma. Such a difference allows tissue type to be characterised according to is diseased state. Furthermore the information can be applied to improve diagnosis. It is proposed that collection and analysis of the scattered x-rays present during a mammographic procedure can supply the additional information and be used to improve the image contrast. The ultimate aim of the project is to improve the specificity of x-ray mammography. (authors) 10 refs., 3 figs.

  8. The value of X-ray CT in orbital fractures

    Energy Technology Data Exchange (ETDEWEB)

    Chung, Myung Hee; Lee, Jae Mun; Kim, Choon Yul; Bahk, Yong Whee [Catholic Medical College, Seoul (Korea, Republic of)

    1986-08-15

    On the pulse from the trauma transiting to posterior side of the orbit, orbital fractures are occurred through the weak point of the orbital wall. Invagination of soft tissue or entrapment of muscles may be associated with orbital fracture. In condition of inaccurate diagnosis, appropriate surgical repairment is impossible and complication such as diplopia or enophthalmia are developed. CT scan is diagnostic procedure which demonstrates accurately the site and state of orbital fracture, and its associated findings. The authors has been studied in 21 orbital CT scan to evaluate the relative value of plain X rays and CT scans in the diagnosis of orbital fractures during the period from January 1982 to September 1985. The conclusions were as follows: 1. Diagnostic rate was 100% by CT, 40% by initial and 80% by retrospective interpretation of conventional X-ray films. 2. Low X-ray diagnostic rate of medical wall fractures (26.7%) was due to thinness of the bone. 3. Medial wall fractures were associated with floor fractures in 46%. 4. Orbital soft tissue injuries and abnormalities of PNS were precisely evaluated by CT scan.

  9. Frameless image registration of X-ray CT and SPECT by volume matching

    International Nuclear Information System (INIS)

    Tanaka, Yuko; Kihara, Tomohiko; Yui, Nobuharu; Kinoshita, Fujimi; Kamimura, Yoshitsugu; Yamada, Yoshifumi.

    1998-01-01

    Image registration of functional (SPECT) and morphological (X-ray CT/MRI) images is studied in order to improve the accuracy and the quantity of the image diagnosis. We have developed a new frameless registration method of X-ray CT and SPECT image using transmission CT image acquired for absorption correction of SPECT images. This is the automated registration method and calculates the transformation matrix between the two coordinate systems of image data by the optimization method. This registration method is based on the similar physical property of X-ray CT and transmission CT image. The three-dimensional overlap of the bone region is used for image matching. We verified by a phantom test that it can provide a good result of within two millimeters error. We also evaluated visually the accuracy of the registration method by the application study of SPECT, X-ray CT, and transmission CT head images. This method can be carried out accurately without any frames. We expect this registration method becomes an efficient tool to improve image diagnosis and medical treatment. (author)

  10. Integrated image presentation of transmission and fluorescent X-ray CT using synchrotron radiation

    Energy Technology Data Exchange (ETDEWEB)

    Zeniya, T.; Takeda, T. E-mail: ttakeda@md.tsukuba.ac.jp; Yu, Q.; Hasegawa, Y.; Hyodo, K.; Yuasa, T.; Hiranaka, Y.; Itai, Y.; Akatsuka, T

    2001-07-21

    We have developed a computed tomography (CT) system with synchrotron radiation (SR) to detect fluorescent X-rays and transmitted X-rays simultaneously. Both SR transmission X-ray CT (SR-TXCT) and SR fluorescent X-ray CT (SR-FXCT) can describe cross-sectional images with high spatial and contrast resolutions as compared to conventional CT. TXCT gives morphological information and FXCT gives functional information of organs. So, superposed display system for SR-FXCT and SR-TXCT images has been developed for clinical diagnosis with higher reliability. Preliminary experiment with brain phantom was carried out and the superposition of both images was performed. The superposed SR-CT image gave us both functional and morphological information easily with high reliability, thus demonstrating the usefulness of this system.

  11. Integrated image presentation of transmission and fluorescent X-ray CT using synchrotron radiation

    Science.gov (United States)

    Zeniya, T.; Takeda, T.; Yu, Q.; Hasegawa, Y.; Hyodo, K.; Yuasa, T.; Hiranaka, Y.; Itai, Y.; Akatsuka, T.

    2001-07-01

    We have developed a computed tomography (CT) system with synchrotron radiation (SR) to detect fluorescent X-rays and transmitted X-rays simultaneously. Both SR transmission X-ray CT (SR-TXCT) and SR fluorescent X-ray CT (SR-FXCT) can describe cross-sectional images with high spatial and contrast resolutions as compared to conventional CT. TXCT gives morphological information and FXCT gives functional information of organs. So, superposed display system for SR-FXCT and SR-TXCT images has been developed for clinical diagnosis with higher reliability. Preliminary experiment with brain phantom was carried out and the superposition of both images was performed. The superposed SR-CT image gave us both functional and morphological information easily with high reliability, thus demonstrating the usefulness of this system.

  12. X-ray CT core imaging of Oman Drilling Project on D/V CHIKYU

    Science.gov (United States)

    Michibayashi, K.; Okazaki, K.; Leong, J. A. M.; Kelemen, P. B.; Johnson, K. T. M.; Greenberger, R. N.; Manning, C. E.; Harris, M.; de Obeso, J. C.; Abe, N.; Hatakeyama, K.; Ildefonse, B.; Takazawa, E.; Teagle, D. A. H.; Coggon, J. A.

    2017-12-01

    We obtained X-ray computed tomography (X-ray CT) images for all cores (GT1A, GT2A, GT3A and BT1A) in Oman Drilling Project Phase 1 (OmanDP cores), since X-ray CT scanning is a routine measurement of the IODP measurement plan onboard Chikyu, which enables the non-destructive observation of the internal structure of core samples. X-ray CT images provide information about chemical compositions and densities of the cores and is useful for assessing sample locations and the quality of the whole-round samples. The X-ray CT scanner (Discovery CT 750HD, GE Medical Systems) on Chikyu scans and reconstructs the image of a 1.4 m section in 10 minutes and produces a series of scan images, each 0.625 mm thick. The X-ray tube (as an X-ray source) and the X-ray detector are installed inside of the gantry at an opposing position to each other. The core sample is scanned in the gantry with the scanning rate of 20 mm/sec. The distribution of attenuation values mapped to an individual slice comprises the raw data that are used for subsequent image processing. Successive two-dimensional (2-D) slices of 512 x 512 pixels yield a representation of attenuation values in three-dimensional (3-D) voxels of 512 x 512 by 1600 in length. Data generated for each core consist of core-axis-normal planes (XY planes) of X-ray attenuation values with dimensions of 512 × 512 pixels in 9 cm × 9 cm cross-section, meaning at the dimensions of a core section, the resolution is 0.176 mm/pixel. X-ray intensity varies as a function of X-ray path length and the linear attenuation coefficient (LAC) of the target material is a function of the chemical composition and density of the target material. The basic measure of attenuation, or radiodensity, is the CT number given in Hounsfield units (HU). CT numbers of air and water are -1000 and 0, respectively. Our preliminary results show that CT numbers of OmanDP cores are well correlated to gamma ray attenuation density (GRA density) as a function of chemical

  13. Monte Carlo evaluation of glandular dose in cone-beam X-ray computed tomography dedicated to the breast: Homogeneous and heterogeneous breast models.

    Science.gov (United States)

    Sarno, Antonio; Mettivier, Giovanni; Tucciariello, Raffaele M; Bliznakova, Kristina; Boone, John M; Sechopoulos, Ioannis; Di Lillo, Francesca; Russo, Paolo

    2018-06-07

    In cone-beam computed tomography dedicated to the breast (BCT), the mean glandular dose (MGD) is the dose metric of reference, evaluated from the measured air kerma by means of normalized glandular dose coefficients (DgN CT ). This work aimed at computing, for a simple breast model, a set of DgN CT values for monoenergetic and polyenergetic X-ray beams, and at validating the results vs. those for patient specific digital phantoms from BCT scans. We developed a Monte Carlo code for calculation of monoenergetic DgN CT coefficients (energy range 4.25-82.25 keV). The pendant breast was modelled as a cylinder of a homogeneous mixture of adipose and glandular tissue with glandular fractions by mass of 0.1%, 14.3%, 25%, 50% or 100%, enveloped by a 1.45 mm-thick skin layer. The breast diameter ranged between 8 cm and 18 cm. Then, polyenergetic DgN CT coefficients were analytically derived for 49-kVp W-anode spectra (half value layer 1.25-1.50 mm Al), as in a commercial BCT scanner. We compared the homogeneous models to 20 digital phantoms produced from classified 3D breast images. Polyenergetic DgN CT resulted 13% lower than most recent published data. The comparison vs. patient specific breast phantoms showed that the homogeneous cylindrical model leads to a DgN CT percentage difference between -15% and +27%, with an average overestimation of 8%. A dataset of monoenergetic and polyenergetic DgN CT coefficients for BCT was provided. Patient specific breast models showed a different volume distribution of glandular dose and determined a DgN CT 8% lower, on average, than homogeneous breast model. Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  14. X-ray CT high-density artefact suppression in cryosurgery

    International Nuclear Information System (INIS)

    Wei Jikun; Sandison, George A; Chen Laigao; Liang Yun; Xu, Lisa X

    2002-01-01

    Advantages of x-ray CT for imaging guidance of cryosurgery include 3D visualization of frozen and unfrozen tissue and calibration of temperature in the tissue water-ice interface (0-10 deg. C) to Hounsfield units. However, use of x-ray CT images and their thermal calibration can be compromised by the cryoprobes generating high-density streak artefacts. A new subtraction technique for artefact suppression is proposed and tested in prostate cryosurgery simulations. By subtracting the measured CT x-ray projection profile without cryoprobes from the profile with cryoprobes plus iceballs, one obtains the combined profile of the cryoprobes and a low value background. Polynomial interpolation to obtain the background profile allows its addition to the original profile without probes. The result may then be fed to a conventional filtered back-projection routine to reconstruct the probe-free image. Finally the cryoprobe pixels in the originally constructed image with probes and iceballs are added back to the probe-free image to get the final artefact-suppressed image. The major advantage of this subtraction technique is that it can successfully suppress the high-density artefacts in bone-abundant body regions such as the pelvis. X-ray CT images of cryoprobe arrays in a homogeneous gelatin phantom and the pelvic region of an anthropomorphic Rando phantom containing a human skeleton were generated. After suppression, cryoprobe metal artefact streaks are reduced and visualization of the positions and dimensions of the cryoprobes are well preserved. (note)

  15. Micron-CT using quasi-monochromatic x-rays produced in micro-PIXE

    International Nuclear Information System (INIS)

    Ishii, K.

    2009-01-01

    In ion-atom collision, characteristic X-rays are intensively produced and can be considered as a monochromatic X-ray source. We apply this feature to X-ray CT. By using micro-beams, cross sectional images can be provided with a spatial resolution of about 1 μm. On the basis of this idea, we developed a micron-CT consisting of a micro-beam system and an X-ray CCD camera. A tube holding samples was rotated by a stepping motor and the transmission images of the sample were taken with characteristic K-X-rays of Ti (4.558 keV) produced by 3 MeV proton micro-beams. After image reconstruction, images of cross sections of small objects were obtained with a spatial resolution of 3 μm. Using an absorption edge, we can identify an element in a sample. It is expected that our micron-CT can provide cross sectional images of in-vivo cellular samples and can be applied to a wide range of researches in biology and medicine. (author)

  16. TU-EF-207-05: Dedicated Cone-beam Breast CT

    International Nuclear Information System (INIS)

    Vedantham, S.

    2015-01-01

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation

  17. TU-EF-207-05: Dedicated Cone-beam Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Vedantham, S. [Univ. of Massachusetts Medical School (United States)

    2015-06-15

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation.

  18. Analogue experiment of the crustal deformation by X-ray CT; X sen CT wo mochiita chikaku no henkei no analogue jikken

    Energy Technology Data Exchange (ETDEWEB)

    Nakajima, Y.; Shi, B; Murakami, Y. [Geological Survey of Japan, Tsukuba (Japan)

    1997-05-27

    This paper reports a predictive result on a deformation experiment on sand beds using an X-ray computerized tomography (CT) equipment. The X-ray CT is a method to perform the following processes: X-rays are irradiated on a sample; decayed intensity data are measured along ray paths of the permeated X-rays; the data are inverted by using the Fourier transform; and spatial distribution of the coefficient of X-ray absorption inside the sample is acquired as a gray scale image. The deformation experiment was carried by putting sand into an acrylic container to a depth of 2 cm and manually pressing a plate in the horizontal direction. Sand blocks make a relative motion along a specific plane when they are deformed, but it was not possible to recognize faults clearly by naked eyes. Upon completion of the deformation, the experimental device was placed on the CT equipment to acquire images of the two-dimensional cross section. The higher the X-ray absorption coefficient, the brighter the picture elements. It can be seen that three inversed faults have been imaged clearly. Therefore, this equipment was verified usable as an effective observation equipment for an analogue deformation experiment. 4 refs., 2 figs.

  19. Evaluation of the BreastSimulator software platform for breast tomography

    Science.gov (United States)

    Mettivier, G.; Bliznakova, K.; Sechopoulos, I.; Boone, J. M.; Di Lillo, F.; Sarno, A.; Castriconi, R.; Russo, P.

    2017-08-01

    The aim of this work was the evaluation of the software BreastSimulator, a breast x-ray imaging simulation software, as a tool for the creation of 3D uncompressed breast digital models and for the simulation and the optimization of computed tomography (CT) scanners dedicated to the breast. Eight 3D digital breast phantoms were created with glandular fractions in the range 10%-35%. The models are characterised by different sizes and modelled realistic anatomical features. X-ray CT projections were simulated for a dedicated cone-beam CT scanner and reconstructed with the FDK algorithm. X-ray projection images were simulated for 5 mono-energetic (27, 32, 35, 43 and 51 keV) and 3 poly-energetic x-ray spectra typically employed in current CT scanners dedicated to the breast (49, 60, or 80 kVp). Clinical CT images acquired from two different clinical breast CT scanners were used for comparison purposes. The quantitative evaluation included calculation of the power-law exponent, β, from simulated and real breast tomograms, based on the power spectrum fitted with a function of the spatial frequency, f, of the form S(f)  =  α/f   β . The breast models were validated by comparison against clinical breast CT and published data. We found that the calculated β coefficients were close to that of clinical CT data from a dedicated breast CT scanner and reported data in the literature. In evaluating the software package BreastSimulator to generate breast models suitable for use with breast CT imaging, we found that the breast phantoms produced with the software tool can reproduce the anatomical structure of real breasts, as evaluated by calculating the β exponent from the power spectral analysis of simulated images. As such, this research tool might contribute considerably to the further development, testing and optimisation of breast CT imaging techniques.

  20. Magnetic resonance imaging of the central nervous system. Comparison with X-ray CT

    Energy Technology Data Exchange (ETDEWEB)

    Kajima, Toshio; Kagawa, Yoshihiro; Katsuta, Shizutomo.

    1987-06-01

    Magnetic resonance imaging (MRI) and X-ray computed tomography (X-ray CT) have been performed in 169 consecutive patients with central nervous system diseases. The findings from the two methods were compared for the capacity to defect lesions. Magnetic resonance imaging was more sensitive than or equivalent to X-ray CT in detecting lesions - especially detecting. Arnold-Chiari malformation, syringomyelia, spinal cord injury, and pituitary adenoma - in 158 patients (94 %). In six patients (10 %), lesion detection was possible only by MRI. Magnetic resonance imaging was inferior to X-ray CT in 11 patients (7 %) in detecting calcified lesions, meningioma, and cavernous hemangioma. (Namekawa, K.).

  1. Registration of SPECT, PET and/or X-ray CT images in patients with lung cancer

    International Nuclear Information System (INIS)

    Uemura, K.; Toyama, H.; Miyamoto, T.; Yoshikawa, K.; Mori, Y.

    2002-01-01

    Aim: In order to evaluate the therapeutic gain of heavy ion therapy performed on patients with lung cancer, the regional pulmonary functions and the amount of radio tracer accumulation to the tumor, we are investigated by using the region of interest based on anatomical information obtained from X-ray CT. There are many registration techniques for brain images, but not so much for the other organ images that we have studied registration of chest SPECT, PET and/or X-ray CT images. Materials and Methods: Perfusion, ventilation and blood pool images with Tc 99m labeled radiopharmaceuticals and SPECT, tumor images with 11 C-methionine and PET and X-ray CT scans were performed on several patients with lung cancer before and after heavy ion therapy. The registrations of SPECT-CT, PET-CT and CT-CT were performed by using AMIR (Automatic Multimodality Image Registration), which was developed by Babak et al. for registration of brain images. In a case of SPECT-CT registration, each of the three functional images was registered to the X-ray CT image, and the accuracy of each registration was compared. In the studies of PET-CT registration, the transmission images and X-ray CT images were registered at first, because the 11 C-methionine PET images bear little resemblance to the underlying anatomical images. Next, the emission images were realigned by using the same registration parameters. The X-ray CT images obtained from a single subject at the different time were registered to the first X-ray CT images, respectively. Results: In the SPECT-CT registration, the blood pool-CT registration is the best among three SPECT images in visual inspection by radiologists. In the PET-CT registration, the Transmission-CT registrations got good results. Therefore, Emission-CT registrations also got good results. In the CT-CT registration, the X-ray CT images obtained from a single subject at the different time were superimposed well each other except for lower lobe. As the results, it was

  2. X-ray CT in the detection of palm weevils

    International Nuclear Information System (INIS)

    Ma, A.K.W.; Alghamdi, A.A.; Spyrou, N.M.

    2012-01-01

    Early detection of the red palm weevils (RPW) is a major challenge in agriculture among all kinds of palm trees due to the nature of the insect and the difficulty to trace them through their life stages associated with the tree life. Many methods have been applied for the weevil detection such as X-ray diffraction techniques, fluoroscopy and ultrasound. On the other hand, the idea of tomography has been used for other purposes such as the determination of the age of the tree and for applied environmental studies. Such technology can also reveal the weevil in principle. In this study, we explore the use of X-ray CT for weevil detection with the Monte Carlo method. A model of the stem of a palm tree is developed for simulations. MCNPX is chosen to carry out the simulations for the radiography tally in the code. The tally records the 2D data of the X-ray beams irradiating the tree model. An iterative reconstruction method for cone beam CT is applied to obtain the 3D slices of the tree model. We are exploring the minimum number of projection angles and the detectability of the weevil. We shall also report the sensitivity of weevil detection using X-ray CT with a large set of simulations with different weevil sizes and tree diameters. (author)

  3. Development and application of an educational 3D X-ray CT instrument

    International Nuclear Information System (INIS)

    Arakawa, Etsuo; Iwami, Ryutaro; Motohisa, Yasuko; Kamezawa, Chika; Kamogawa, Masashi; Voegeli, Wolfgang

    2016-01-01

    A three-dimensional (3D) X-ray computed tomography (CT) instrument for radiation education was developed. The structure of the instrument is such that the main parts, i.e. the X-ray source, specimen rotation stage, and two-dimensional detector can be easily observed. An experiment using a fruit of green pepper as a specimen was performed. CT images and intermediate steps for obtaining them, i.e. radiographs, sinograms after Radon transform, and real and imaginary parts of Fourier components in reciprocal space during inverse Radon transform are shown. We propose that these images will help students to understand the principle and mechanism of X-ray CT instruments visually. (author)

  4. Estimation of effective dose from limited cone beam X-ray CT examination

    Energy Technology Data Exchange (ETDEWEB)

    Iwai, Kazuo; Arai, Yoshinori; Hashimoto, Koji [Nihon Univ., Tokyo (Japan). School of Dentistry; Nishizawa, Kanae

    2000-12-01

    The limited cone beam X-ray CT (Ortho-CT) was developed on the basis of multi-functional panoramic apparatus, SCANORA (Soredex Co. Helsinki Finland). The imaging intensifier (I.I.) was built in this apparatus as a X-ray detection device instead of X-ray film. The signal provided from I.I. was converted from analog into digital by an analog-digital converter and image reconstitution was done as a three-directional image of the dimensions 3.8 cm of width, 3.0 cm height and 3.8 cm depth with the personal computer. The 3DX Multi image micro CT'' (3DX) was developed along similar lines by MORITA Co., Ltd. (Kyoto, JAPAN). In this study, the stochastic effect on organ and tissue caused by examinations using Ortho-CT and 3DX was measured. The effective dose was estimated according to the recommendation of ICRP60 and was compared with those of panoramic radiography and computed tomography. The irradiation conditions were as follows: 85 kV, 10 mA with the filtration of 3 mmAl and added 1 mmCu for Ortho-CT, and 80 kV, 2 mA and the filtration of 3.1 mmAL for 3DX. The measurement of organ and tissue dose was performed using an anthropomorphic Rando woman phantom (Alderson Research Laboratories Co., Stanfora, CN), as well as by using two different type of thermoluminescent dosimeter (TLD); Panasonic UD-170A (BeO) and UD-110S (CaSO{sub 4}: Tm). The UD-170A was for dose measurement of the inner useful X-ray beams, while the UD-110S was for outer beams. The measured organ and tissue were those recommended with ICRP60 (gonad, breast, bone marrow, lung, thyroid gland, esophagus, stomach, colon, liver, bladder, skin, brain, thymus, adrenal, kidney, spleen, pancrease, upper large intestine, uterus, eyes and major salivary gland). The imaging by Orhto-CT was made in the left maxillary 1st molar, left mandibular 1st molar and temporomandibular joint. 3DX measurement was made in the maxillary incisor region and middle ear regions other than the regions mentioned above. The skin

  5. Library based x-ray scatter correction for dedicated cone beam breast CT

    International Nuclear Information System (INIS)

    Shi, Linxi; Zhu, Lei; Vedantham, Srinivasan; Karellas, Andrew

    2016-01-01

    Purpose: The image quality of dedicated cone beam breast CT (CBBCT) is limited by substantial scatter contamination, resulting in cupping artifacts and contrast-loss in reconstructed images. Such effects obscure the visibility of soft-tissue lesions and calcifications, which hinders breast cancer detection and diagnosis. In this work, we propose a library-based software approach to suppress scatter on CBBCT images with high efficiency, accuracy, and reliability. Methods: The authors precompute a scatter library on simplified breast models with different sizes using the GEANT4-based Monte Carlo (MC) toolkit. The breast is approximated as a semiellipsoid with homogeneous glandular/adipose tissue mixture. For scatter correction on real clinical data, the authors estimate the breast size from a first-pass breast CT reconstruction and then select the corresponding scatter distribution from the library. The selected scatter distribution from simplified breast models is spatially translated to match the projection data from the clinical scan and is subtracted from the measured projection for effective scatter correction. The method performance was evaluated using 15 sets of patient data, with a wide range of breast sizes representing about 95% of general population. Spatial nonuniformity (SNU) and contrast to signal deviation ratio (CDR) were used as metrics for evaluation. Results: Since the time-consuming MC simulation for library generation is precomputed, the authors’ method efficiently corrects for scatter with minimal processing time. Furthermore, the authors find that a scatter library on a simple breast model with only one input parameter, i.e., the breast diameter, sufficiently guarantees improvements in SNU and CDR. For the 15 clinical datasets, the authors’ method reduces the average SNU from 7.14% to 2.47% in coronal views and from 10.14% to 3.02% in sagittal views. On average, the CDR is improved by a factor of 1.49 in coronal views and 2.12 in sagittal

  6. Library based x-ray scatter correction for dedicated cone beam breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Shi, Linxi; Zhu, Lei, E-mail: leizhu@gatech.edu [Nuclear and Radiological Engineering and Medical Physics Programs, The George W. Woodruff School of Mechanical Engineering, Georgia Institute of Technology, Atlanta, Georgia 30332 (United States); Vedantham, Srinivasan; Karellas, Andrew [Department of Radiology, University of Massachusetts Medical School, Worcester, Massachusetts 01655 (United States)

    2016-08-15

    Purpose: The image quality of dedicated cone beam breast CT (CBBCT) is limited by substantial scatter contamination, resulting in cupping artifacts and contrast-loss in reconstructed images. Such effects obscure the visibility of soft-tissue lesions and calcifications, which hinders breast cancer detection and diagnosis. In this work, we propose a library-based software approach to suppress scatter on CBBCT images with high efficiency, accuracy, and reliability. Methods: The authors precompute a scatter library on simplified breast models with different sizes using the GEANT4-based Monte Carlo (MC) toolkit. The breast is approximated as a semiellipsoid with homogeneous glandular/adipose tissue mixture. For scatter correction on real clinical data, the authors estimate the breast size from a first-pass breast CT reconstruction and then select the corresponding scatter distribution from the library. The selected scatter distribution from simplified breast models is spatially translated to match the projection data from the clinical scan and is subtracted from the measured projection for effective scatter correction. The method performance was evaluated using 15 sets of patient data, with a wide range of breast sizes representing about 95% of general population. Spatial nonuniformity (SNU) and contrast to signal deviation ratio (CDR) were used as metrics for evaluation. Results: Since the time-consuming MC simulation for library generation is precomputed, the authors’ method efficiently corrects for scatter with minimal processing time. Furthermore, the authors find that a scatter library on a simple breast model with only one input parameter, i.e., the breast diameter, sufficiently guarantees improvements in SNU and CDR. For the 15 clinical datasets, the authors’ method reduces the average SNU from 7.14% to 2.47% in coronal views and from 10.14% to 3.02% in sagittal views. On average, the CDR is improved by a factor of 1.49 in coronal views and 2.12 in sagittal

  7. Gas scintillation glass GEM detector for high-resolution X-ray imaging and CT

    Energy Technology Data Exchange (ETDEWEB)

    Fujiwara, T., E-mail: fujiwara-t@aist.go.jp [Research Institute for Measurement and Analytical Instrumentation, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Ibaraki 305-8568 (Japan); Mitsuya, Y. [Nuclear Professional School, The University of Tokyo, Tokai, Naka, Ibaraki 319-1188 (Japan); Fushie, T. [Radiment Lab. Inc., Setagaya, Tokyo 156-0044 (Japan); Murata, K.; Kawamura, A.; Koishikawa, A. [XIT Co., Naruse, Machida, Tokyo 194-0045 (Japan); Toyokawa, H. [Research Institute for Measurement and Analytical Instrumentation, National Institute of Advanced Industrial Science and Technology (AIST), Tsukuba, Ibaraki 305-8568 (Japan); Takahashi, H. [Institute of Engineering Innovation, School of Engineering, The University of Tokyo, Bunkyo, Tokyo 113-8654 (Japan)

    2017-04-01

    A high-spatial-resolution X-ray-imaging gaseous detector has been developed with a single high-gas-gain glass gas electron multiplier (G-GEM), scintillation gas, and optical camera. High-resolution X-ray imaging of soft elements is performed with a spatial resolution of 281 µm rms and an effective area of 100×100 mm. In addition, high-resolution X-ray 3D computed tomography (CT) is successfully demonstrated with the gaseous detector. It shows high sensitivity to low-energy X-rays, which results in high-contrast radiographs of objects containing elements with low atomic numbers. In addition, the high yield of scintillation light enables fast X-ray imaging, which is an advantage for constructing CT images with low-energy X-rays.

  8. Power spectrum analysis of the x-ray scatter signal in mammography and breast tomosynthesis projections.

    Science.gov (United States)

    Sechopoulos, Ioannis; Bliznakova, Kristina; Fei, Baowei

    2013-10-01

    To analyze the frequency domain characteristics of the signal in mammography images and breast tomosynthesis projections with patient tissue texture due to detected scattered x-rays. Acquisitions of x-ray projection images of 19 different patient breasts were simulated using previously acquired volumetric patient images. Acquisition of these images was performed with a dedicated breast CT prototype system, and the images were classified into voxels representing skin, adipose, and glandular tissue with a previously validated automated algorithm. The classified three dimensional images then underwent simulated mechanical compression representing that which is performed during acquisition of mammography and breast tomosynthesis images. The acquisition of projection images of each patient breast was simulated using Monte Carlo methods with each simulation resulting in two images: one of the primary (non-scattered) signal and one of the scatter signal. To analyze the scatter signal for both mammography and breast tomosynthesis, two projections images of each patient breast were simulated, one with the x-ray source positioned at 0° (mammography and central tomosynthesis projection) and at 30° (wide tomosynthesis projection). The noise power spectra (NPS) for both the scatter signal alone and the total signal (primary + scatter) for all images were obtained and the combined results of all patients analyzed. The total NPS was fit to the expected power-law relationship NPS(f) = k/f β and the results were compared with those previously published on the power spectrum characteristics of mammographic texture. The scatter signal alone was analyzed qualitatively and a power-law fit was also performed. The mammography and tomosynthesis projections of three patient breasts were too small to analyze, so a total of 16 patient breasts were analyzed. The values of β for the total signal of the 0° projections agreed well with previously published results. As expected, the scatter

  9. TU-F-18C-03: X-Ray Scatter Correction in Breast CT: Advances and Patient Testing

    International Nuclear Information System (INIS)

    Ramamurthy, S; Sechopoulos, I

    2014-01-01

    Purpose: To further develop and perform patient testing of an x-ray scatter correction algorithm for dedicated breast computed tomography (BCT). Methods: A previously proposed algorithm for x-ray scatter signal reduction in BCT imaging was modified and tested with a phantom and on patients. A wireless electronic positioner system was designed and added to the BCT system that positions a tungsten plate in and out of the x-ray beam. The interpolation used by the algorithm was replaced with a radial basis function-based algorithm, with automated exclusion of non-valid sampled points due to patient motion or other factors. A 3D adaptive noise reduction filter was also introduced to reduce the impact of scatter quantum noise post-reconstruction. The impact on image quality of the improved algorithm was evaluated using a breast phantom and seven patient breasts, using quantitative metrics such signal difference (SD) and signal difference-to-noise ratios (SDNR) and qualitatively using image profiles. Results: The improvements in the algorithm resulted in a more robust interpolation step, with no introduction of image artifacts, especially at the imaged object boundaries, which was an issue in the previous implementation. Qualitative evaluation of the reconstructed slices and corresponding profiles show excellent homogeneity of both the background and the higher density features throughout the whole imaged object, as well as increased accuracy in the Hounsfield Units (HU) values of the tissues. Profiles also demonstrate substantial increase in both SD and SDNR between glandular and adipose regions compared to both the uncorrected and system-corrected images. Conclusion: The improved scatter correction algorithm can be reliably used during patient BCT acquisitions with no introduction of artifacts, resulting in substantial improvement in image quality. Its impact on actual clinical performance needs to be evaluated in the future. Research Agreement, Koning Corp., Hologic

  10. Normalized glandular dose (DgN) coefficients for flat-panel CT breast imaging

    International Nuclear Information System (INIS)

    Thacker, Samta C; Glick, Stephen J

    2004-01-01

    The development of new digital mammography techniques such as dual-energy imaging, tomosynthesis and CT breast imaging will require investigation of optimal camera design parameters and optimal imaging acquisition parameters. In optimizing these acquisition protocols and imaging systems it is important to have knowledge of the radiation dose to the breast. This study presents a methodology for estimating the normalized glandular dose to the uncompressed breast using the geometry proposed for flat-panel CT breast imaging. The simulation uses the GEANT 3 Monte Carlo code to model x-ray transport and absorption within the breast phantom. The Monte Carlo software was validated for breast dosimetry by comparing results of the normalized glandular dose (DgN) values of the compressed breast to those reported in the literature. The normalized glandular dose was then estimated for a range of breast diameters from 10 cm to 18 cm using an uncompressed breast model with a homogeneous composition of adipose and glandular tissue, and for monoenergetic x-rays from 10 keV to 120 keV. These data were fit providing expressions for the normalized glandular dose. Using these expressions for the DgN coefficients and input variables such as the diameter, height and composition of the breast phantom, the mean glandular dose for any spectra can be estimated. A computer program to provide normalized glandular dose values has been made available online. In addition, figures displaying energy deposition maps are presented to better understand the spatial distribution of dose in CT breast imaging

  11. Application of small-angle X-ray scattering for differentiation among breast tumors

    International Nuclear Information System (INIS)

    Changizi, V.; Kheradmand, A. Arab; Oghabian, M.A.

    2008-01-01

    Small-angle X-ray scattering (SAXS) is an X-ray diffraction-based technique where a narrow collimated beam of X-rays is focused onto a sample and the scattered X-rays recorded by a detector. The pattern of the scattered X-rays carries information on the molecular structure of the material. As breast cancer is the most widespread cancer in women and differentiation among its tumors is important, this project compared the results of coherent X-ray scattering measurements obtained from benign and malignant breast tissues. The energy-dispersive method with a setup including X-ray tube, primary collimator, sample holder, secondary collimator and high-purity germanium (HpGe) detector was used. One hundred thirty-one breast-tissue samples, including normal, fibrocystic changes and carcinoma, were studied at the 6 deg scattering angle. Diffraction profiles (corrected scattered intensity versus momentum transfer) of normal, fibrocystic changes and carcinoma were obtained. These profiles showed a few peak positions for adipose (1.15 ± 0.06 nm -1 ), mixed normal (1.15 ± 0.06 nm -1 and 1.4 ± 0.04 nm -1 ), fibrocystic changes (1.46 ± 0.05 nm -1 and 1.74 ± 0.04 nm -1 ) and carcinoma (1.55 ± 0.04 nm -1 , 1.73 ± 0.06 nm -1 , 1.85 ± 0.05 nm -1 ). We were able to differentiate between normal, fibrocystic changes (benign) and carcinoma (malignant) breast tissues by SAXS. However, we were unable to differentiate between different types of carcinoma. (author)

  12. Application of small-angle X-ray scattering for differentiation among breast tumors

    Directory of Open Access Journals (Sweden)

    Changizi V

    2008-01-01

    Full Text Available Small-angle X-ray scattering (SAXS is an X-ray diffraction-based technique where a narrow collimated beam of X-rays is focused onto a sample and the scattered X-rays recorded by a detector. The pattern of the scattered X-rays carries information on the molecular structure of the material. As breast cancer is the most widespread cancer in women and differentiation among its tumors is important, this project compared the results of coherent X-ray scattering measurements obtained from benign and malignant breast tissues. The energy-dispersive method with a setup including X-ray tube, primary collimator, sample holder, secondary collimator and high-purity germanium (HpGe detector was used. One hundred thirty-one breast-tissue samples, including normal, fibrocystic changes and carcinoma, were studied at the 6° scattering angle. Diffraction profiles (corrected scattered intensity versus momentum transfer of normal, fibrocystic changes and carcinoma were obtained. These profiles showed a few peak positions for adipose (1.15 ± 0.06 nm -1 , mixed normal (1.15 ± 0.06 nm -1 and 1.4 ± 0.04 nm -1 , fibrocystic changes (1.46 ± 0.05 nm -1 and 1.74 ± 0.04 nm -1 and carcinoma (1.55 ± 0.04 nm -1 , 1.73 ± 0.06 nm -1 , 1.85 ± 0.05 nm -1 . We were able to differentiate between normal, fibrocystic changes (benign and carcinoma (malignant breast tissues by SAXS. However, we were unable to differentiate between different types of carcinoma.

  13. X-ray, CT and MRI findings of synovial tuberculosis in joints

    International Nuclear Information System (INIS)

    Yu Jinghong; Tao Meili; You Zhuangzhi; Yu Huazhi

    2006-01-01

    Objective: To analyze the X-ray, CT and MRI findings of synovial tuberculosis, and to evaluate the role of MRI in diagnosing synovial tuberculosis. Methods: Fourteen cases of synovial tuberculosis comfirmed by operation and pathology were retrospectively analyzed and summarized. All patients were examined by MRI and X-ray, and CT scans were performed in 3 cases. Results: X-ray showed joint swelling (8 cases), articular space narrowing (7 cases), marginal joint erosions (4 cases), and periarticular osteoporosis (9 cases). The joint swelling was detected on CT in all 3 cases, and bony erosion and speckled sequestra were seen in 2 cases. MRI in all of patients showed joint swelling and synovial proliferation in different drgees, demonstrated as heterogeneously low signal on T 1 WI and slight high signal (7 cases) and obvious high signal (6 cases) on T 2 WI, and diffuse synovial proliferation was demonstrated as massive and nodular signal in 8 cases. Joint effusion was present in 7 cases as low signal on T 1 WI and high signal on T 2 WI. Osseous erosion lesions were seen in 7 cases, and intra-articular cartilage thinned, partly or mostly disappeared in 11 cases. Periarticular bone marrow edema was found in 7 cases. Conclusion: MRI was superior to X-ray and CT in the diagnosis and differential diagnosis of synovial tuberculosis. (authors)

  14. A new cone-beam X-ray CT system with a reduced size planar detector

    International Nuclear Information System (INIS)

    Li Liang; Chen Zhiqiang; Zhang Li; Xing Yuxiang; Kang Kejun

    2006-01-01

    In a traditional cone-beam CT system, the cost of product and computation is very high. The authors propose a transversely truncated cone-beam X-ray CT system with a reduced size detector positioned off-center, in which X-ray beams only cover half of the object. The reduced detector size cuts the cost and the X-ray dose of the CT system. The existing CT reconstruction algorithms are not directly applicable in this new CT system. Hence, the authors develop a BPF-type direct backprojection algorithm. Different from the traditional rebinding methods, our algorithm directly backprojects the pretreated projection data without rebinding. This makes the algorithm compact and computationally more efficient. Finally, some numerical simulations and practical experiments are done to validate the proposed algorithm. (authors)

  15. Positron emission CT and X-ray CT findings in chronic obstructive pulmonary diseases

    Energy Technology Data Exchange (ETDEWEB)

    Sato, Yoshikazu; Murata, Kiyoshi; Ito, Harumi; Senda, Michio; Yonekura, Yoshiharu; Konishi, Junji; Nishimura, Koichi; Izumi, Takahide; Oshima, Shunsaku

    1987-08-01

    Positron emission CT and X-ray CT were performed in fifteen patients with emphysema confirmed SAB and twelve patients with clinical DPB. In patients with emphysema, 20 of 36 areas showed a central pattern and their perfusion scintigrams showed stripe-signs. On the other hand, the patients with DPB showed outer layer progression of the disease.

  16. X-ray and CT findings of soft tissue and bone infections secondary to acquired immunodeficiency syndrome

    International Nuclear Information System (INIS)

    Jiang Songfeng; Liu Jinxin; Chen Bihua; Zhang Lieguang; Gan Qingxin; Huang Deyang

    2011-01-01

    Objective: To summarize X-ray and CT findings of soft tissue and bone infections secondary to acquired immunodeficiency syndrome (AIDS). Methods: The data of X-ray and CT findings of soft tissue and bone infections in 18 patients with AIDS were retrospectively collected and analyzed. Results: Of 18 patients with AIDS, the CT features of soft tissue demonstrated that subcutaneous patchy high density in 1 case which considered as cellulitis, round low density lesions with ring enhancement in 6 cases which considered as soft tissue abscesses, heterogeneous density lesions with peripheral enhancement in 1 case which considered as pyomyositis. Of 18 patients with AIDS, septic arthritis was found in 4 cases involving knee lesion in 3 cases and hip lesion. In the 4 case, the X-ray films showed bony destruction in 2 cases and the CT showed bone destruction in 3 cases and arthroedema in 4 cases. Of 18 patients with AIDS, osteomyelitis was found in 9 cases of which tuberculosis was considered in. 8 cases and vertebral involvement in 6 cases. In the 9 cases, the X-ray films and CT displayed bony destruction, hyperostosis, small sequestra, and intervertebral space narrowing. Of 18 patients with AIDS, costal lesions were found in 3 cases in which the CT showed expandable bony destruction. Of 18 patients with AIDS, ilium and sacroiliac joint lesions were found in 1 case in which the X-ray films and CT showed bony destruction, sequestra, and joint widening. Of 18 patients with AIDS, chronic pyogenic osteomyelitis of femur was found in 1 case in which the X-ray films showed bony destruction, hyperostosis osteosclerosis, and periosteal reaction. Conclusion: The X-ray and CT features of soft tissue and bone infections secondary to AIDS are characterized. The X-ray and CT are useful tools to early diagnose soft tissue and bone infections secondary to AIDS. (authors)

  17. Grating Oriented Line-Wise Filtration (GOLF) for Dual-Energy X-ray CT

    Science.gov (United States)

    Xi, Yan; Cong, Wenxiang; Harrison, Daniel; Wang, Ge

    2017-12-01

    In medical X-ray Computed Tomography (CT), the use of two distinct X-ray source spectra (energies) allows dose-reduction and material discrimination relative to that achieved with only one source spectrum. Existing dual-energy CT methods include source kVp-switching, double-layer detection, dual-source gantry, and two-pass scanning. Each method suffers either from strong spectral correlation or patient-motion artifacts. To simultaneously address these problems, we propose to improve CT data acquisition with the Grating Oriented Line-wise Filtration (GOLF) method, a novel X-ray filter that is placed between the source and patient. GOLF uses a combination of absorption and filtering gratings that are moved relative to each other and in synchronization with the X-ray tube kVp-switching process and/or the detector view-sampling process. Simulation results show that GOLF can improve the spectral performance of kVp-switching to match that of dual-source CT while avoiding patient motion artifacts and dual imaging chains. Although significant flux is absorbed by this pre-patient filter, the proposed GOLF method is a novel path for cost-effectively extracting dual-energy or multi-energy data and reducing radiation dose with or without kVp switching.

  18. A dedicated breast-PET/CT scanner: Evaluation of basic performance characteristics.

    Science.gov (United States)

    Raylman, Raymond R; Van Kampen, Will; Stolin, Alexander V; Gong, Wenbo; Jaliparthi, Gangadhar; Martone, Peter F; Smith, Mark F; Sarment, David; Clinthorne, Neal H; Perna, Mark

    2018-04-01

    Application of advanced imaging techniques, such as PET and x ray CT, can potentially improve detection of breast cancer. Unfortunately, both modalities have challenges in the detection of some lesions. The combination of the two techniques, however, could potentially lead to an overall improvement in diagnostic breast imaging. The purpose of this investigation is to test the basic performance of a new dedicated breast-PET/CT. The PET component consists of a rotating pair of detectors. Its performance was evaluated using the NEMA NU4-2008 protocols. The CT component utilizes a pulsed x ray source and flat panel detector mounted on the same gantry as the PET scanner. Its performance was assessed using specialized phantoms. The radiation dose to a breast during CT imaging was explored by the measurement of free-in-air kerma and air kerma measured at the center of a 16 cm-diameter PMMA cylinder. Finally, the combined capabilities of the system were demonstrated by imaging of a micro-hot-rod phantom. Overall, performance of the PET component is comparable to many pre-clinical and other dedicated breast-PET scanners. Its spatial resolution is 2.2 mm, 5 mm from the center of the scanner using images created with the single-sliced-filtered-backprojection algorithm. Peak NECR is 24.6 kcps; peak sensitivity is 1.36%; the scatter fraction is 27%. Spatial resolution of the CT scanner is 1.1 lp/mm at 10% MTF. The free-in-air kerma is 2.33 mGy, while the PMMA-air kerma is 1.24 mGy. Finally, combined imaging of a micro-hot-rod phantom illustrated the potential utility of the dual-modality images produced by the system. The basic performance characteristics of a new dedicated breast-PET/CT scanner are good, demonstrating that its performance is similar to current dedicated PET and CT scanners. The potential value of this system is the capability to produce combined duality-modality images that could improve detection of breast disease. The next stage in development of this system

  19. WE-EF-207-06: Dedicated Cone-Beam Breast CT with Laterally-Shifted Detector: Monte Carlo Evaluation of X-Ray Scatter Distribution and Scatter-To-Primary Ratio

    International Nuclear Information System (INIS)

    Shi, L; Vedantham, S; Karellas, A

    2015-01-01

    Purpose: To determine the spatial distribution of x-ray scatter and scatter-to-primary ratio (SPR) in projections during cone-beam breast CT (CBBCT) with laterally-shifted detector that results in coronal (fan-angle) truncation. Methods: We hypothesized that CBBCT with coronal truncation would lower SPR due to reduction in irradiated breast volume, and that the location of maximum x-ray scatter fluence (scatter-peak) in the detector plane can be determined from the ratio of irradiated-to-total breast volume, breast dimensions and system geometry. Monte Carlo simulations (GEANT4) reflecting a prototype CBBCT system were used to record the position-dependent primary and scatter x-ray photon fluence incident on the detector without coronal truncation (full fan-angle, 2f=24-degrees) and with coronal truncation (fan-angle, f+ f=12+2.7-degrees). Semi-ellipsoidal breasts (10/14/18-cm diameter, chest-wall to nipple length: 0.75xdiameter, 2%/14%/100% fibroglandular content) aligned with the axis-of-rotation (AOR) were modeled. Mono-energy photons were simulated and weighted for 2 spectra (49kVp, 1.4-mm Al HVL; 60kVp, 3.76-mm Al HVL). In addition to SPR, the scatter maps were analyzed to identify the location of the scatter-peak. Results: For CBBCT without fan-angle truncation, the scatter-peaks were aligned with the projection of the AOR onto the detector for all breasts. With truncated fan-beam, the scatter-peaks were laterally-shifted from the projection of the AOR along the fan-angle direction by 14/38/70-pixels for 10/14/18-cm diameter breasts. The corresponding theoretical shifts were 14.8/39.7/68-pixels (p=0.47, 2-tailed paired-ratio t-test). Along the cone-angle, the shift in scatter-peaks between truncated and full-fan angle CBBCT were 2/2/4 -pixels for 10/14/18-cm diameter breasts. CBBCT with fan-angle truncation reduced SPR by 14/22/28% for 10/14/18-cm diameter breasts. 60kVp reduced SPR by 21–25% compared to 49kVp. Peak SPR for CBBCT with fan-angle truncation

  20. X-ray scattering for the characterization of lyophilized breast tissue samples

    International Nuclear Information System (INIS)

    Elshemey, Wael M.; Mohamed, Fayrouz S.; Khater, Ibrahim M.

    2013-01-01

    This work investigates the possibility of characterizing breast cancer by measuring the X-ray scattering profiles of lyophilized excised breast tissue samples. Since X-ray scattering from water-rich tissue is dominated by scattering from water, the removal of water by lyophilization would enhance the characterization process. In the present study, X-ray scattering profiles of 22 normal, 22 malignant and 10 benign breast tissue samples are measured. The cut-offs of scatter diagrams, sensitivity, specificity and diagnostic accuracy of three characterization parameters (full width at half maximum (FWHM) for the peak at 1.1 nm −1 , area under curve (AUC), and ratio of 1st to 2nd scattering peak intensities (I 1 /I 2 %)) are calculated and compared to the data from non-lyophilized samples. Results show increased sensitivity (up to 100%) of the present data on lyophilized breast tissue samples compared to previously reported data for non-lyophilized samples while the specificity (up to 95.4%), diagnostic accuracy (up to 95.4%) and receiver operating characteristic (ROC) curve values (up to 0.9979) for both sets of data are comparable. The present study shows significant differences between normal samples and each of malignant and benign samples. Only subtle differences exist between malignant and benign lyophilized breast tissue samples where FWHM=0.7±0.1 and 0.8±0.3, AUC=1.3±0.2 and 1.4±0.2 and I 1 /I 2 %=44.9±11.0 and 52.4±7.6 for malignant and benign samples respectively. - Highlights: • X-ray scattering profiles of breast tissue samples are acquired. • Three X-ray profile characterization parameters are calculated. • The cut-offs, sensitivity, specificity and diagnostic accuracy are calculated. • They are compared to the data from non-lyophilized samples. • Results show increased sensitivity in case of lyophilized samples

  1. TU-F-18C-07: Hardware Advances for MTF Improvement in Dedicated Breast CT

    International Nuclear Information System (INIS)

    Gazi, P; Burkett, G; Yang, K; Boone, J

    2014-01-01

    Purpose: In this study, we have designed and implemented a prototype dedicated breast CT system (bCT) to improve the spatial resolution characteristics, in order to improve detection of micro-calcifications. Methods: A 10.8 kW water-cooled, tungsten anode x-ray tube, running up to 240 mA at 60 kV, coupled with an x-ray generator specifically designed for this application, and 0.3 mm of added copper filter was used to generate x-ray pulses. A CsI CMOS flat panel detector with a pixel pitch of 0.075 mm in native binning mode was used. The system geometry was designed in a way to achieve an FOV on par with similar bCT prototypes, resulting in a magnification factor of 1.39. A 0.013 mm tungsten wire was used to generate point spread functions. Multiple scans were performed with different numbers of projections, different reconstruction kernel sizes and different reconstruction filters to study the effects of each parameter on MTF. The resulting MTFs were then evaluated quantitatively using the generated PFSs. Duplicate scans with the same parameters were performed on two other dedicated breast CT systems to compare the performance of the new prototype. Results: The results of the MTF experiments demonstrate a significant improvement in the spatial resolution characteristics. In the new prototype, using the pulsed x-ray source results in a restoration of the azimuthal MTF degradation, due to motion blurring previously seen in other bCT systems. Moreover, employing the higher resolution x-ray detector considerably improves the MTF. The MTF at 10% of the new system is at 3.5 1/mm, a factor of 4.36 greater than an earlier bCT scanner. Conclusion: The MTF analysis of the new prototype bCT shows that using the new hardware and control results in a significant improvement in visualization of finer detail. This suggests that the visualization of micro-calcifications will be significantly improved

  2. Scaling-law for the energy dependence of anatomic power spectrum in dedicated breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Vedantham, Srinivasan; Shi, Linxi; Glick, Stephen J.; Karellas, Andrew [Department of Radiology, University of Massachusetts Medical School, Worcester, Massachusetts 01655 (United States)

    2013-01-15

    Purpose: To determine the x-ray photon energy dependence of the anatomic power spectrum of the breast when imaged with dedicated breast computed tomography (CT). Methods: A theoretical framework for scaling the empirically determined anatomic power spectrum at one x-ray photon energy to that at any given x-ray photon energy when imaged with dedicated breast CT was developed. Theory predicted that when the anatomic power spectrum is fitted with a power curve of the form k f{sup -{beta}}, where k and {beta} are fit coefficients and f is spatial frequency, the exponent {beta} would be independent of x-ray photon energy (E), and the amplitude k scales with the square of the difference in energy-dependent linear attenuation coefficients of fibroglandular and adipose tissues. Twenty mastectomy specimens based numerical phantoms that were previously imaged with a benchtop flat-panel cone-beam CT system were converted to 3D distribution of glandular weight fraction (f{sub g}) and were used to verify the theoretical findings. The 3D power spectrum was computed in terms of f{sub g} and after converting to linear attenuation coefficients at monoenergetic x-ray photon energies of 20-80 keV in 5 keV intervals. The 1D power spectra along the axes were extracted and fitted with a power curve of the form k f{sup -{beta}}. The energy dependence of k and {beta} were analyzed. Results: For the 20 mastectomy specimen based numerical phantoms used in the study, the exponent {beta} was found to be in the range of 2.34-2.42, depending on the axis of measurement. Numerical simulations agreed with the theoretical predictions that for a power-law anatomic spectrum of the form k f{sup -{beta}}, {beta} was independent of E and k(E) =k{sub 1}[{mu}{sub g}(E) -{mu}{sub a}(E)]{sup 2}, where k{sub 1} is a constant, and {mu}{sub g}(E) and {mu}{sub a}(E) represent the energy-dependent linear attenuation coefficients of fibroglandular and adipose tissues, respectively. Conclusions: Numerical

  3. Using x-ray mammograms to assist in microwave breast image interpretation.

    Science.gov (United States)

    Curtis, Charlotte; Frayne, Richard; Fear, Elise

    2012-01-01

    Current clinical breast imaging modalities include ultrasound, magnetic resonance (MR) imaging, and the ubiquitous X-ray mammography. Microwave imaging, which takes advantage of differing electromagnetic properties to obtain image contrast, shows potential as a complementary imaging technique. As an emerging modality, interpretation of 3D microwave images poses a significant challenge. MR images are often used to assist in this task, and X-ray mammograms are readily available. However, X-ray mammograms provide 2D images of a breast under compression, resulting in significant geometric distortion. This paper presents a method to estimate the 3D shape of the breast and locations of regions of interest from standard clinical mammograms. The technique was developed using MR images as the reference 3D shape with the future intention of using microwave images. Twelve breast shapes were estimated and compared to ground truth MR images, resulting in a skin surface estimation accurate to within an average Euclidean distance of 10 mm. The 3D locations of regions of interest were estimated to be within the same clinical area of the breast as corresponding regions seen on MR imaging. These results encourage investigation into the use of mammography as a source of information to assist with microwave image interpretation as well as validation of microwave imaging techniques.

  4. SCT-4800T whole body X-ray CT scanner

    International Nuclear Information System (INIS)

    Okumura, Yoshitaka; Sato, Yukio; Kuwahara, Hiroshi

    1994-01-01

    A whole body X-ray CT scanner, the SCT-4800T (trade name: INTELLECT series), has been developed. This system is the first CT scanner that is combined with general radiographic functions. The general radiographic functions include a patient couch with film casette and several tube support systems along with the CT scanner. This newly designed CT scanner also features a compact and light-weight gantry with a 700 mm diameter apperture and user-friendly operater's console. The SCT-4800T brings a new level of patient and operator comfort to the emergency radiology examination site. (author)

  5. TH-AB-209-07: High Resolution X-Ray-Induced Acoustic Computed Tomography

    Energy Technology Data Exchange (ETDEWEB)

    Xiang, L; Tang, S [University of Oklahoma, Norman, OK (United States); Ahmad, M [Stanford University, Palo Alto, CA (United States); Xing, L [Stanford University School of Medicine, Stanford, CA (United States)

    2016-06-15

    Purpose: X-ray radiographic absorption imaging is an invaluable tool in medical diagnostics, biology and materials science. However, the use of conventional CT is limited by two factors: the detection sensitivity to weak absorption material and the radiation dose from CT scanning. The purpose of this study is to explore X-ray induced acoustic computed tomography (XACT), a new imaging modality, which combines X-ray absorption contrast and high ultrasonic resolution to address these challenges. Methods: First, theoretical models was built to analyze the XACT sensitivity to X-ray absorption and calculate the minimal radiation dose in XACT imaging. Then, an XACT system comprised of an ultrashort X-ray pulse, a low noise ultrasound detector and a signal acquisition system was built to evaluate the X-ray induced acoustic signal generation. A piece of chicken bone and a phantom with two golden fiducial markers were exposed to 270 kVp X-ray source with 60 ns exposure time, and the X-ray induced acoustic signal was received by a 2.25MHz ultrasound transducer in 200 positions. XACT images were reconstructed by a filtered back-projection algorithm. Results: The theoretical analysis shows that X-ray induced acoustic signals have 100% relative sensitivity to X-ray absorption, but not to X-ray scattering. Applying this innovative technology to breast imaging, we can reduce radiation dose by a factor of 50 compared with newly FDA approved breast CT. The reconstructed images of chicken bone and golden fiducial marker phantom reveal that the spatial resolution of the built XACT system is 350µm. Conclusion: In XACT, the imaging sensitivity to X-ray absorption is improved and the imaging dose is dramatically reduced by using ultrashort pulsed X-ray. Taking advantage of the high ultrasonic resolution, we can also perform 3D imaging with a single X-ray pulse. This new modality has the potential to revolutionize x-ray imaging applications in medicine and biology.

  6. MRI of atlantoaxial subluxation; Correlating with plain X-ray films and CT findings

    Energy Technology Data Exchange (ETDEWEB)

    Saikawa, Yuko; Nishi, Naoko; Saitoh, Yoko; Akimura, Rumiko; Sasaki, Taisuke; Yodono, Hiraku; Takekawa, Shoichi; Harata, Seikou; Sannohe, Akio (Hirosaki Univ., Aomori (Japan). School of Medicine)

    1991-04-01

    Twenty-three patients with atlantoaxial subluxation (14 with rheumatoid arthritis, one each with rheumatoid arthritis and Arnold-Chiari malformation, os odontoideum, Klippel-Feil syndrome, trauma, and 5 with unknown causes) were evaluated with MRI. We used 0.5 T MRI unit (RESONA; Yokogawa Medical Systems, Japan) and 1.5 T MRI unit (SMT 150; Shimazu, Japan) with head or flexible coils. We compared the usefulness of MRI with those of plain X-ray films and CT regarding several points. MRI provided better image of the soft tissue mass around the odontoid process, compression of cord or subarachnoid space than plain X-ray films and CT. Atlanto-odontoid distance on MRI is nearer to that on plain X-ray films than CT. MRI is useful in analyzing the anatomic details such as transverse ligament, alar ligament, tectorial membrane and thickened synovium. Both MRI and CT provided detailed bony changes. High correlation was observed between MRI grading of cord compression and the degree of myelopathy. (author).

  7. Head injury in childhood: comparison of sonography with the conventional X-ray and CT

    International Nuclear Information System (INIS)

    Steiner, S.; Riebel, T.; Nazarenko, O.; Bassir, C.; Steger, W.; Vogl, T.; Felix, R.

    1996-01-01

    Purpose: The purpose of our study was to compare the value of ultrasound, conventional X-ray diagnosis and CT in detecting skull fractures and intracranial haemorrhage in children suffering from a head injury. Material and methods: We examined 210 children who had a head injury. In all cases the calvarium was investigated by ultrasound using a 7.0 MHz linear transducer. In children with an open fontanel (n=190) the cerebrum was screened additionally by ultrasound following a standard protocol. The sonographic findings were correlated to the X-ray examination (n=21) and CT (n=13). Results: Ultrasound enabled diagnosis of linear calvarial fractures (n=29), depressed fratures (n=6) and intracranial haemorrhage (n=8). X-Ray and XT examination confirmed the diagnosis of linear calvarial fractures in 16 cases, of depressed fractures in 6 cases. CT confirmed the sonographic diagnosis of intracranial haemorrhage in 8 cases. Conclusion: Ultrasound as a primary method can replace the conventional X-ray in detecting calvarial fracture and posttraumatic sequelae. Additional CT examination depends on the sonographic and neurological status. (orig.) [de

  8. Differential diagnosis of benign and malignant breast tumors by high frequency molybdenum-target X-ray photography

    International Nuclear Information System (INIS)

    Mai Yuanqi; Wang Maosheng; Huang Jian; Cui Guoru; Liang Zhicong; Lu Yingying

    2008-01-01

    Objective: To explore the X-ray Image of benign and malignant breast lesions (tumors) in order to improve their differcatial diagnostic level. Methods: X-ray image changes of 63 malignant breast neoplasms were described by the mammography and in comparision with those of 43 benign masses. Results: The accordance percentages between the X-ray and histological examinations for the benign and malignant neoplasms were shown as 85% and 90.6% respectively. Spiculated mass, calcification granules in clusters and other images were found to be indication of benign or malignant breast lesion. Conclusion: The High Frequency Molybdenum-target X-ray Photography can provide effective imaging data for diagnosis and distinguish between the benign and malignant breast lesions. (authors)

  9. Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetry

    International Nuclear Information System (INIS)

    Jirasek, A; Hilts, M; McAuley, K B

    2010-01-01

    A primary limitation of current x-ray CT polymer gel dosimetry is the low contrast, and hence poor dose resolution, of dose images produced by the system. The low contrast is largely due to the low-dose sensitivity of current formulations of polymer gel for x-ray CT imaging. This study reports on the investigation of new dosimeter formulations with improved dose sensitivity for x-ray CT polymer gel dosimetry. We incorporate an isopropanol co-solvent into an N-isopropylacrylamide-based gel formulation in order to increase the total monomer/crosslinker concentration (%T) within the formulation. It is shown that gels of high %T exhibit enhanced dose sensitivity and dose resolutions over traditional formulations. The gels are shown to be temporally stable and reproducible. A single formulation (16%T) is used to demonstrate the capabilities of the x-ray CT polymer gel dosimetry system in measuring known dose distributions. A 1 L gel volume is exposed to three separate irradiations: a single-field percent depth dose, a two-field 'cross' and a three-field 'test case'. The first two irradiations are used to generate a dose calibration curve by which images are calibrated. The calibrated images are compared with treatment planning predictions and it is shown that the x-ray CT polymer gel dosimetry system is capable of capturing spatial and dose information accurately. The proposed new gel formulation is shown to be sensitive, stable and to improve the dose resolution over current formulations so as to provide a feasible gel for clinical applications of x-ray CT polymer gel dosimetry.

  10. Fatigue damage observed non-destructively in fibre composite coupon test specimens by X-ray CT

    DEFF Research Database (Denmark)

    Jespersen, Kristine Munk; Mikkelsen, Lars Pilgaard

    2016-01-01

    This study presents a method for monitoring the 3D fatigue damage progression on a micro-structural level in a glass fibre/polymer coupon test specimen by means of laboratory X-ray Computed Tomography (CT). A modified mount and holder made for the standard test samples to fit into the X-ray CT...... scanner along with a tension clamp solution is presented. Initially, the same location of the test specimen is inspected by ex-situ X-ray CT during the fatigue loading history, which shows the damage progression on a micro-structural level. The openings of individual uni-directional (UD) fibre fractures...

  11. Using X-Ray Mammograms to Assist in Microwave Breast Image Interpretation

    Directory of Open Access Journals (Sweden)

    Charlotte Curtis

    2012-01-01

    Full Text Available Current clinical breast imaging modalities include ultrasound, magnetic resonance (MR imaging, and the ubiquitous X-ray mammography. Microwave imaging, which takes advantage of differing electromagnetic properties to obtain image contrast, shows potential as a complementary imaging technique. As an emerging modality, interpretation of 3D microwave images poses a significant challenge. MR images are often used to assist in this task, and X-ray mammograms are readily available. However, X-ray mammograms provide 2D images of a breast under compression, resulting in significant geometric distortion. This paper presents a method to estimate the 3D shape of the breast and locations of regions of interest from standard clinical mammograms. The technique was developed using MR images as the reference 3D shape with the future intention of using microwave images. Twelve breast shapes were estimated and compared to ground truth MR images, resulting in a skin surface estimation accurate to within an average Euclidean distance of 10 mm. The 3D locations of regions of interest were estimated to be within the same clinical area of the breast as corresponding regions seen on MR imaging. These results encourage investigation into the use of mammography as a source of information to assist with microwave image interpretation as well as validation of microwave imaging techniques.

  12. Dosimetric comparison between CT and X-ray simulation of radiotherapy for lung cancer

    International Nuclear Information System (INIS)

    Kali Ayguli, Zhang Jinrong; Wang Juwu; Ge Feng; Wang Haifeng; Xu Suling

    2007-01-01

    Objective: To compare radiotherapy plan of conventional X-ray simulation with CT simulation by 3D-TPS for lung cancer. Methods: Thirty-three patients were allotted to receive both conventional X-ray simulation and CT simulation in the same treatment position. 3D-TPS was used to design 4-field conventional plan of X-ray simulation (RT), 4-field two dimensional plan(2D)and three dimensional conformal radiation plan(3DCRT) of CT simulation for all patients. The total dose was 50 Gy. Dose volume histogram(DVH) was applied to evaluate the difference of target coverage, dose distribution and normal tissue protection among the three plans. Results: 3DCRT and 2D based on CT simulation were superior to RT in the target coverage, target conformity index (TCI) and target homogeneity (TH) (P 20 , V 30 and mean lung dose were similar among 3DCRT, 2D and RT plans. Moreover, the maximum doses of spinal cord were significantly different among the three plans. No statistical differences of doses to 30% of the heart and esophagus volume among the three plans were observed. Conclusions: There is significantly better tumour volume coverage in CT simulation when compared with X-ray conventional simulation. Target volume delineation by CT simulation is improved significantly. The dose distribution is improved by using three dimensional treatment planning system. 3DCRT plan is superior to 2D plans in target conformity index and target homogeneity. Doses delivered to organs surrounding the target such as lung and heart were reduced significantly in 3DCRT. (authors)

  13. Interferometric phase-contrast X-ray CT imaging of VX2 rabbit cancer at 35keV X-ray energy

    Science.gov (United States)

    Takeda, Tohoru; Wu, Jin; Tsuchiya, Yoshinori; Yoneyama, Akio; Lwin, Thet-Thet; Hyodo, Kazuyuki; Itai, Yuji

    2004-05-01

    Imaging of large objects at 17.7-keV low x-ray energy causes huge x-ray exposure to the objects even using interferometric phase-contrast x-ray CT (PCCT). Thus, we tried to obtain PCCT images at high x-ray energy of 35keV and examined the image quality using a formalin-fixed VX2 rabbit cancer specimen with 15-mm in diameter. The PCCT system consisted of an asymmetrically cut silicon (220) crystal, a monolithic x-ray interferometer, a phase-shifter, an object cell and an x-ray CCD camera. The PCCT at 35 keV clearly visualized various inner structures of VX2 rabbit cancer such as necrosis, cancer, the surrounding tumor vessels, and normal liver tissue. Besides, image-contrast was not degraded significantly. These results suggest that the PCCT at 35 KeV is sufficient to clearly depict the histopathological morphology of VX2 rabbit cancer specimen.

  14. Polymer gel dosimeters with enhanced sensitivity for use in x-ray CT polymer gel dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Jirasek, A [Department of Physics and Astronomy, University of Victoria, Victoria BC V8W 3P6 (Canada); Hilts, M [Medical Physics, BC Cancer Agency-Vancouver Island Centre, Victoria BC V6R 2B6 (Canada); McAuley, K B, E-mail: jirasek@uvic.c [Department of Chemical Engineering, Queens University, Kingston, ON K7 L 3N6 (Canada)

    2010-09-21

    A primary limitation of current x-ray CT polymer gel dosimetry is the low contrast, and hence poor dose resolution, of dose images produced by the system. The low contrast is largely due to the low-dose sensitivity of current formulations of polymer gel for x-ray CT imaging. This study reports on the investigation of new dosimeter formulations with improved dose sensitivity for x-ray CT polymer gel dosimetry. We incorporate an isopropanol co-solvent into an N-isopropylacrylamide-based gel formulation in order to increase the total monomer/crosslinker concentration (%T) within the formulation. It is shown that gels of high %T exhibit enhanced dose sensitivity and dose resolutions over traditional formulations. The gels are shown to be temporally stable and reproducible. A single formulation (16%T) is used to demonstrate the capabilities of the x-ray CT polymer gel dosimetry system in measuring known dose distributions. A 1 L gel volume is exposed to three separate irradiations: a single-field percent depth dose, a two-field 'cross' and a three-field 'test case'. The first two irradiations are used to generate a dose calibration curve by which images are calibrated. The calibrated images are compared with treatment planning predictions and it is shown that the x-ray CT polymer gel dosimetry system is capable of capturing spatial and dose information accurately. The proposed new gel formulation is shown to be sensitive, stable and to improve the dose resolution over current formulations so as to provide a feasible gel for clinical applications of x-ray CT polymer gel dosimetry.

  15. Early experience with a picture archiving and communication system for X-ray CT

    International Nuclear Information System (INIS)

    Nakano, Y.; Komori, M.; Minato, K.

    1987-01-01

    A network using fiberoptic ring architecture has been implemented to connect an x-ray CT unit with an image work station and optical disk storage unit. The work station has three noninterlaced cathode-ray tube (CRT) monitors of 1,024 X 1,024 pixels. The optical disk system stores 32 optical disk cartridges (2.6 Gbytes per cartridge). A general-purpose interface bus is used for communication between the CT unit and the network interface unit. Transfer time from CT to interface unit is 40 kbyte/second, 5 seconds per image. On-line data acquisition, long-term archiving, and CT interpretation using CRTs have been tested and the system has been working for 3 months in a clinical environment

  16. Analysis of local concentration in stomach fold pattern by using abdominal X-ray CT image

    International Nuclear Information System (INIS)

    Watanabe, Shigeto; Hasegawa, Jun-ichi; Mekada, Yoshito; Mori, Kensaku; Nawano, Shigeru

    2004-01-01

    The paper presents a method for analysis of stomach folds pattern in abdominal X-ray CT images. Many stomach cancers have local concentration of folds at cancer lesions. Therefore, in stomach diagnosis, it is very important to evaluate quantitatively such fold concentration. In this paper, a method for calculation of concentration index for line figure on a curved surface is proposed first. Then, using this method a processing procedure for detection of stomach cancer from an abdominal X-ray CT image is developed. In the experiment using practical X-ray CT images, it is shown that by the proposed procedure, higher values of concentration index are obtained at cancer area. (author)

  17. Patient size and x-ray transmission in body CT.

    Science.gov (United States)

    Ogden, Kent; Huda, Walter; Scalzetti, Ernest M; Roskopf, Marsha L

    2004-04-01

    Physical characteristics were obtained for 196 patients undergoing chest and abdomen computed tomography (CT) examinations. Computed tomography sections for these patients having no evident pathology were analyzed to determine patient dimensions (AP and lateral), together with the average attenuation coefficient. Patient weights ranged from approximately 3 kg to about 120 kg. For chest CT, the mean Hounsfield unit (HU) fell from about -120 HU for newborns to about -300 HU for adults. For abdominal CT, the mean HU for children and normal-sized adults was about 20 HU, but decreased to below -50 HU for adults weighing more than 100 kg. The effective photon energy and percent energy fluence transmitted through a given patient size and composition was calculated for representative x-ray spectra at 80, 100, 120, and 140 kV tube potentials. A 70-kg adult scanned at 120 kVp transmits 2.6% of the energy fluence for chest and 0.7% for abdomen CT examinations. Reducing the patient size to 10 kg increases transmission by an order of magnitude. For 70 kg patients, effective energies in body CT range from approximately 50 keV at 80 kVp to approximately 67 keV at 140 kVp; increasing patient size from 10 to 120 kg resulted in an increase in effective photon energy of approximately 4 keV. The x-ray transmission data and effective photon energy data can be used to determine CT image noise and image contrast, respectively, and information on patient size and composition can be used to determine patient doses.

  18. High-energy x-ray CT and its application for digital engineering

    International Nuclear Information System (INIS)

    Kamimura, H.; Sadaoka, N.

    2005-01-01

    A high-energy x-ray computed tomography system and x-ray CT data handling software have been developed for digital engineering; internal dimension measurement, density analysis, actual and designed shape comparison, STL file generation, and support for reverse engineering and rapid prototyping. The system is designed to collect accurate images in short scanning time (10 s per section) using a MeV-energy electron linear accelerator and highly sensitive semiconductor detectors in order to scan large objects made of aluminum and/or iron. An excellent environment in digital engineering is provided by the software products; 'StereoCooker' for 3D bitmap CAD (rendering, feature extraction, dimensional measurement, and shape comparison, etc.), 'FeatureMaker' for translating bitmap CT data to CAD data including feature information, and 'Wingware' for realizing an Windows PC cluster system 'WINGluster' to apply CT data analysis. (author)

  19. Noise variance analysis using a flat panel x-ray detector: A method for additive noise assessment with application to breast CT applications

    Energy Technology Data Exchange (ETDEWEB)

    Yang Kai; Huang, Shih-Ying; Packard, Nathan J.; Boone, John M. [Department of Radiology, University of California, Davis Medical Center, 4860 Y Street, Suite 3100 Ellison Building, Sacramento, California 95817 (United States); Department of Radiology, University of California, Davis Medical Center, 4860 Y Street, Suite 3100 Ellison Building, Sacramento, California 95817 (United States) and Department of Biomedical Engineering, University of California, Davis, Davis, California, 95616 (United States)

    2010-07-15

    Purpose: A simplified linear model approach was proposed to accurately model the response of a flat panel detector used for breast CT (bCT). Methods: Individual detector pixel mean and variance were measured from bCT projection images acquired both in air and with a polyethylene cylinder, with the detector operating in both fixed low gain and dynamic gain mode. Once the coefficients of the linear model are determined, the fractional additive noise can be used as a quantitative metric to evaluate the system's efficiency in utilizing x-ray photons, including the performance of different gain modes of the detector. Results: Fractional additive noise increases as the object thickness increases or as the radiation dose to the detector decreases. For bCT scan techniques on the UC Davis prototype scanner (80 kVp, 500 views total, 30 frames/s), in the low gain mode, additive noise contributes 21% of the total pixel noise variance for a 10 cm object and 44% for a 17 cm object. With the dynamic gain mode, additive noise only represents approximately 2.6% of the total pixel noise variance for a 10 cm object and 7.3% for a 17 cm object. Conclusions: The existence of the signal-independent additive noise is the primary cause for a quadratic relationship between bCT noise variance and the inverse of radiation dose at the detector. With the knowledge of the additive noise contribution to experimentally acquired images, system modifications can be made to reduce the impact of additive noise and improve the quantum noise efficiency of the bCT system.

  20. Bone Mechanical Strength Estimation from Micro X-ray CT Image

    National Research Council Canada - National Science Library

    Matani, A

    2001-01-01

    ... (Bone Mineral Density), an index to evaluate the mechanical strength of the bone, does not always reflect the strength, On the other hand, micro X-ray CT has revealed the inner structure of bone, Under such circumstances...

  1. MRI features of myositis ossificans with X-ray and CT findings

    International Nuclear Information System (INIS)

    Gu Xiang; Bai Rongjie; Qu Hui; Cheng Xiaoguang; Li Yuang

    2009-01-01

    Objective: To investigate the MRI features of myositis ossificans, and to address the correlation with X-ray radiography and CT findings. Methods: X-ray films, CT and MRI of 36 documented cases of myositis ossificans were retrospectively analyzed, and the literatures were reviewed. Results: Of the 36 cases, 4 cases occurred in the elbow joint, 4 in the shoulder joint, 15 in the hip joint, 6 in the tibiofibula, 5 in the femur, 1 in the metatarsal bones, and 1 in the ilium, respectively. Irregular patchy or lamellar high density calcification or ossification could be seen in the soft tissue parenchym on X-ray films and CT scan. Cortical bone integrity was preserved in diaphysis. CT enhanced scan showed that the swollen parenchyma was not enhanced and there was no parenchyma mass. On the early and middle stages, MR T 1 WI and T 2 WI showed slice-shaped low signal in the peripheral parenchyma, but patchy high signal was found around the low signal on T 2 WI. STIR showed mixed high and low signals in the swollen parenchyma with unclear demarcation. The lesions showed low signal on MR T 1 WI and T 2 WI in the late stage, and there was no edema in peripheral parenehyma. MRI enhanced scan found that the swollen parenchyma showed no enhancement in all stages. Conclusions: The imaging features of myositis ossificans have some characteristics. Misdiagnosis could be avoided when the disease was evaluated with the course. (authors)

  2. A deep convolutional neural network using directional wavelets for low-dose X-ray CT reconstruction.

    Science.gov (United States)

    Kang, Eunhee; Min, Junhong; Ye, Jong Chul

    2017-10-01

    Due to the potential risk of inducing cancer, radiation exposure by X-ray CT devices should be reduced for routine patient scanning. However, in low-dose X-ray CT, severe artifacts typically occur due to photon starvation, beam hardening, and other causes, all of which decrease the reliability of the diagnosis. Thus, a high-quality reconstruction method from low-dose X-ray CT data has become a major research topic in the CT community. Conventional model-based de-noising approaches are, however, computationally very expensive, and image-domain de-noising approaches cannot readily remove CT-specific noise patterns. To tackle these problems, we want to develop a new low-dose X-ray CT algorithm based on a deep-learning approach. We propose an algorithm which uses a deep convolutional neural network (CNN) which is applied to the wavelet transform coefficients of low-dose CT images. More specifically, using a directional wavelet transform to extract the directional component of artifacts and exploit the intra- and inter- band correlations, our deep network can effectively suppress CT-specific noise. In addition, our CNN is designed with a residual learning architecture for faster network training and better performance. Experimental results confirm that the proposed algorithm effectively removes complex noise patterns from CT images derived from a reduced X-ray dose. In addition, we show that the wavelet-domain CNN is efficient when used to remove noise from low-dose CT compared to existing approaches. Our results were rigorously evaluated by several radiologists at the Mayo Clinic and won second place at the 2016 "Low-Dose CT Grand Challenge." To the best of our knowledge, this work is the first deep-learning architecture for low-dose CT reconstruction which has been rigorously evaluated and proven to be effective. In addition, the proposed algorithm, in contrast to existing model-based iterative reconstruction (MBIR) methods, has considerable potential to benefit from

  3. Non-destructive test for irradiated fuels using X-ray CT system in hot-laboratory

    International Nuclear Information System (INIS)

    Kim, Heemoon; Kim, Gil-Soo; Yoo, Boung-Ok; Tahk, Young-Wook; Cho, Moon-Sung; Ahn, Sang-Bok

    2015-01-01

    To inspect inside of irradiated fuel rod for PIE in hotcell, neutron beam and X-ray have been used. Many hot laboratories in the world have shown the results for NDT by 2-D film data. Currently, computed image processing technology instead of film has been developed and CT was applied to the X-ray and neutron beam system. In this trend, our facility needed to set up X-ray system for irradiated fuel inspection and installed in hotcell with consideration of radiation damage. In this study, X-ray system was tested to be operated with radioactive samples and was performed to inspect fuel rods and observe internal damage and dimensional change. 450kV X-ray CT system was installed in hotcell with modification and tested to check image resolution and radiation damage. The image data were analyzed by 3-D computer software. 8 fuel plates and VHTR rods were inspected and measured internal shape and dimension

  4. Combined X-ray CT and mass spectrometry for biomedical imaging applications

    Science.gov (United States)

    Schioppa, E., Jr.; Ellis, S.; Bruinen, A. L.; Visser, J.; Heeren, R. M. A.; Uher, J.; Koffeman, E.

    2014-04-01

    Imaging technologies play a key role in many branches of science, especially in biology and medicine. They provide an invaluable insight into both internal structure and processes within a broad range of samples. There are many techniques that allow one to obtain images of an object. Different techniques are based on the analysis of a particular sample property by means of a dedicated imaging system, and as such, each imaging modality provides the researcher with different information. The use of multimodal imaging (imaging with several different techniques) can provide additional and complementary information that is not possible when employing a single imaging technique alone. In this study, we present for the first time a multi-modal imaging technique where X-ray computerized tomography (CT) is combined with mass spectrometry imaging (MSI). While X-ray CT provides 3-dimensional information regarding the internal structure of the sample based on X-ray absorption coefficients, MSI of thin sections acquired from the same sample allows the spatial distribution of many elements/molecules, each distinguished by its unique mass-to-charge ratio (m/z), to be determined within a single measurement and with a spatial resolution as low as 1 μm or even less. The aim of the work is to demonstrate how molecular information from MSI can be spatially correlated with 3D structural information acquired from X-ray CT. In these experiments, frozen samples are imaged in an X-ray CT setup using Medipix based detectors equipped with a CO2 cooled sample holder. Single projections are pre-processed before tomographic reconstruction using a signal-to-thickness calibration. In the second step, the object is sliced into thin sections (circa 20 μm) that are then imaged using both matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) and secondary ion (SIMS) mass spectrometry, where the spatial distribution of specific molecules within the sample is determined. The

  5. Study on Compression Induced Contrast in X-ray Mammograms Using Breast Mimicking Phantoms

    Directory of Open Access Journals (Sweden)

    A. B. M. Aowlad Hossain

    2015-09-01

    Full Text Available X-ray mammography is commonly used to scan cancer or tumors in breast using low dose x-rays. But mammograms suffer from low contrast problem. The breast is compressed in mammography to reduce x-ray scattering effects. As tumors are stiffer than normal tissues, they undergo smaller deformation under compression. Therefore, image intensity at tumor region may change less than the background tissues. In this study, we try to find out compression induced contrast from multiple mammographic images of tumorous breast phantoms taken with different compressions. This is an extended work of our previous simulation study with experiment and more analysis. We have used FEM models for synthetic phantom and constructed a phantom using agar and n-propanol for simulation and experiment. The x-ray images of deformed phantoms have been obtained under three compression steps and a non-rigid registration technique has been applied to register these images. It is noticeably observed that the image intensity changes at tumor are less than those at surrounding which induce a detectable contrast. Addition of this compression induced contrast to the simulated and experimental images has improved their original contrast by a factor of about 1.4

  6. Cone-beam volume CT breast imaging: Feasibility study

    International Nuclear Information System (INIS)

    Chen Biao; Ning Ruola

    2002-01-01

    X-ray projection mammography, using a film/screen combination, or digital techniques, has proven to be the most effective imaging modality currently available for early detection of breast cancer. However, the inherent superimposition of structures makes a small carcinoma (a few millimeters in size) difficult to detect when it is occult or in dense breasts, leading to a high false-positive biopsy rate. Cone-beam x-ray-projection-based volume imaging using flat panel detectors (FPDs) may allow obtaining three-dimensional breast images, resulting in more accurate diagnosis of structures and patterns of lesions while eliminating the hard compression of breasts. This article presents a novel cone-beam volume computed tomographic breast imaging (CBVCTBI) technique based on the above techniques. Through a variety of computer simulations, the key issues of the system and imaging techniques were addressed, including the x-ray imaging geometry and corresponding reconstruction algorithms, x-ray characteristics of breast tissue and lesions, x-ray setting techniques, the absorbed dose estimation, and the quantitative effect of x-ray scattering on image quality. The preliminary simulation results support the proposed CVBCTBI modality for breast imaging in respect to its feasibility and practicability. The absorbed dose level is comparable to that of current mammography and will not be a prominent problem for this imaging technique. Compared to conventional mammography, the proposed imaging technique with isotropic spatial resolution will potentially provide significantly better low-contrast detectability of breast tumors and more accurate location of breast lesions

  7. A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging

    International Nuclear Information System (INIS)

    Gong Xing; Glick, Stephen J.; Liu, Bob; Vedula, Aruna A.; Thacker, Samta

    2006-01-01

    Although conventional mammography is currently the best modality to detect early breast cancer, it is limited in that the recorded image represents the superposition of a three-dimensional (3D) object onto a 2D plane. Recently, two promising approaches for 3D volumetric breast imaging have been proposed, breast tomosynthesis (BT) and CT breast imaging (CTBI). To investigate possible improvements in lesion detection accuracy with either breast tomosynthesis or CT breast imaging as compared to digital mammography (DM), a computer simulation study was conducted using simulated lesions embedded into a structured 3D breast model. The computer simulation realistically modeled x-ray transport through a breast model, as well as the signal and noise propagation through a CsI based flat-panel imager. Polyenergetic x-ray spectra of Mo/Mo 28 kVp for digital mammography, Mo/Rh 28 kVp for BT, and W/Ce 50 kVp for CTBI were modeled. For the CTBI simulation, the intensity of the x-ray spectra for each projection view was determined so as to provide a total average glandular dose of 4 mGy, which is approximately equivalent to that given in conventional two-view screening mammography. The same total dose was modeled for both the DM and BT simulations. Irregular lesions were simulated by using a stochastic growth algorithm providing lesions with an effective diameter of 5 mm. Breast tissue was simulated by generating an ensemble of backgrounds with a power law spectrum, with the composition of 50% fibroglandular and 50% adipose tissue. To evaluate lesion detection accuracy, a receiver operating characteristic (ROC) study was performed with five observers reading an ensemble of images for each case. The average area under the ROC curves (A z ) was 0.76 for DM, 0.93 for BT, and 0.94 for CTBI. Results indicated that for the same dose, a 5 mm lesion embedded in a structured breast phantom was detected by the two volumetric breast imaging systems, BT and CTBI, with statistically

  8. Methods of X-ray CT image reconstruction from few projections

    International Nuclear Information System (INIS)

    Wang, H.

    2011-01-01

    To improve the safety (low dose) and the productivity (fast acquisition) of a X-ray CT system, we want to reconstruct a high quality image from a small number of projections. The classical reconstruction algorithms generally fail since the reconstruction procedure is unstable and suffers from artifacts. A new approach based on the recently developed 'Compressed Sensing' (CS) theory assumes that the unknown image is in some sense 'sparse' or 'compressible', and the reconstruction is formulated through a non linear optimization problem (TV/l1 minimization) by enhancing the sparsity. Using the pixel (or voxel in 3D) as basis, to apply the CS framework in CT one usually needs a 'sparsifying' transform, and combines it with the 'X-ray projector' which applies on the pixel image. In this thesis, we have adapted a 'CT-friendly' radial basis of Gaussian family called 'blob' to the CS-CT framework. The blob has better space-frequency localization properties than the pixel, and many operations, such as the X-ray transform, can be evaluated analytically and are highly parallelizable (on GPU platform). Compared to the classical Kaisser-Bessel blob, the new basis has a multi-scale structure: an image is the sum of dilated and translated radial Mexican hat functions. The typical medical objects are compressible under this basis, so the sparse representation system used in the ordinary CS algorithms is no more needed. 2D simulations show that the existing TV and l1 algorithms are more efficient and the reconstructions have better visual quality than the equivalent approach based on the pixel or wavelet basis. The new approach has also been validated on 2D experimental data, where we have observed that in general the number of projections can be reduced to about 50%, without compromising the image quality. (author) [fr

  9. Breast neoplasms in women treated with x-rays for acute postpartum mastitis

    International Nuclear Information System (INIS)

    Shore, R.E.; Hempelmann, L.H.; Kowaluk, E.; Mansur, P.S.; Pasternack, B.S.; Albert, R.E.; Haughie, G.E.

    1977-01-01

    Breast cancer has been studied by mail survey up to 34 years in 571 of 606 women treated with x-rays for acute postpartum mastitis. The incidence of neoplasms was compared with that of three nonirradiated control groups--nonirradiated sisters of the treated women, women with acute postpartum mastitis not treated with X-rays, and their nonirradiated sisters. For the irradiated group, with a mean dose of 247 rads to both breasts, the overall relative risk of breast cancer was 2.2 for years 10 to 34 post irradiation and 3.6 for 20 to 34. The dose response for malignant and benign breast neoplasms was compatible with a linear fit. For comparable total doses, fractionation of exposure did not reduce carcinogenic action. Women over age 30 years at radiation treatment had as great an excess risk of breast cancer as did younger women. The overall excess risk of developing breast cancer was about 8 to 10 cases per million women per rad per year, an increase of about 0.5% per rad

  10. Effects of X-Ray Dose On Rhizosphere Studies Using X-Ray Computed Tomography

    Science.gov (United States)

    Zappala, Susan; Helliwell, Jonathan R.; Tracy, Saoirse R.; Mairhofer, Stefan; Sturrock, Craig J.; Pridmore, Tony; Bennett, Malcolm; Mooney, Sacha J.

    2013-01-01

    X-ray Computed Tomography (CT) is a non-destructive imaging technique originally designed for diagnostic medicine, which was adopted for rhizosphere and soil science applications in the early 1980s. X-ray CT enables researchers to simultaneously visualise and quantify the heterogeneous soil matrix of mineral grains, organic matter, air-filled pores and water-filled pores. Additionally, X-ray CT allows visualisation of plant roots in situ without the need for traditional invasive methods such as root washing. However, one routinely unreported aspect of X-ray CT is the potential effect of X-ray dose on the soil-borne microorganisms and plants in rhizosphere investigations. Here we aimed to i) highlight the need for more consistent reporting of X-ray CT parameters for dose to sample, ii) to provide an overview of previously reported impacts of X-rays on soil microorganisms and plant roots and iii) present new data investigating the response of plant roots and microbial communities to X-ray exposure. Fewer than 5% of the 126 publications included in the literature review contained sufficient information to calculate dose and only 2.4% of the publications explicitly state an estimate of dose received by each sample. We conducted a study involving rice roots growing in soil, observing no significant difference between the numbers of root tips, root volume and total root length in scanned versus unscanned samples. In parallel, a soil microbe experiment scanning samples over a total of 24 weeks observed no significant difference between the scanned and unscanned microbial biomass values. We conclude from the literature review and our own experiments that X-ray CT does not impact plant growth or soil microbial populations when employing a low level of dose (<30 Gy). However, the call for higher throughput X-ray CT means that doses that biological samples receive are likely to increase and thus should be closely monitored. PMID:23840640

  11. Adaptive wiener filter based on Gaussian mixture distribution model for denoising chest X-ray CT image

    International Nuclear Information System (INIS)

    Tabuchi, Motohiro; Yamane, Nobumoto; Morikawa, Yoshitaka

    2008-01-01

    In recent decades, X-ray CT imaging has become more important as a result of its high-resolution performance. However, it is well known that the X-ray dose is insufficient in the techniques that use low-dose imaging in health screening or thin-slice imaging in work-up. Therefore, the degradation of CT images caused by the streak artifact frequently becomes problematic. In this study, we applied a Wiener filter (WF) using the universal Gaussian mixture distribution model (UNI-GMM) as a statistical model to remove streak artifact. In designing the WF, it is necessary to estimate the statistical model and the precise co-variances of the original image. In the proposed method, we obtained a variety of chest X-ray CT images using a phantom simulating a chest organ, and we estimated the statistical information using the images for training. The results of simulation showed that it is possible to fit the UNI-GMM to the chest X-ray CT images and reduce the specific noise. (author)

  12. Low-dose X-ray CT reconstruction via dictionary learning.

    Science.gov (United States)

    Xu, Qiong; Yu, Hengyong; Mou, Xuanqin; Zhang, Lei; Hsieh, Jiang; Wang, Ge

    2012-09-01

    Although diagnostic medical imaging provides enormous benefits in the early detection and accuracy diagnosis of various diseases, there are growing concerns on the potential side effect of radiation induced genetic, cancerous and other diseases. How to reduce radiation dose while maintaining the diagnostic performance is a major challenge in the computed tomography (CT) field. Inspired by the compressive sensing theory, the sparse constraint in terms of total variation (TV) minimization has already led to promising results for low-dose CT reconstruction. Compared to the discrete gradient transform used in the TV method, dictionary learning is proven to be an effective way for sparse representation. On the other hand, it is important to consider the statistical property of projection data in the low-dose CT case. Recently, we have developed a dictionary learning based approach for low-dose X-ray CT. In this paper, we present this method in detail and evaluate it in experiments. In our method, the sparse constraint in terms of a redundant dictionary is incorporated into an objective function in a statistical iterative reconstruction framework. The dictionary can be either predetermined before an image reconstruction task or adaptively defined during the reconstruction process. An alternating minimization scheme is developed to minimize the objective function. Our approach is evaluated with low-dose X-ray projections collected in animal and human CT studies, and the improvement associated with dictionary learning is quantified relative to filtered backprojection and TV-based reconstructions. The results show that the proposed approach might produce better images with lower noise and more detailed structural features in our selected cases. However, there is no proof that this is true for all kinds of structures.

  13. Multisource inverse-geometry CT. Part II. X-ray source design and prototype

    Energy Technology Data Exchange (ETDEWEB)

    Neculaes, V. Bogdan, E-mail: neculaes@ge.com; Caiafa, Antonio; Cao, Yang; De Man, Bruno; Edic, Peter M.; Frutschy, Kristopher; Gunturi, Satish; Inzinna, Lou; Reynolds, Joseph; Vermilyea, Mark; Wagner, David; Zhang, Xi; Zou, Yun [GE Global Research, Niskayuna, New York 12309 (United States); Pelc, Norbert J. [Department of Radiology, Stanford University, Stanford, California 94305 (United States); Lounsberry, Brian [Healthcare Science Technology, GE Healthcare, West Milwaukee, Wisconsin 53219 (United States)

    2016-08-15

    Purpose: This paper summarizes the development of a high-power distributed x-ray source, or “multisource,” designed for inverse-geometry computed tomography (CT) applications [see B. De Man et al., “Multisource inverse-geometry CT. Part I. System concept and development,” Med. Phys. 43, 4607–4616 (2016)]. The paper presents the evolution of the source architecture, component design (anode, emitter, beam optics, control electronics, high voltage insulator), and experimental validation. Methods: Dispenser cathode emitters were chosen as electron sources. A modular design was adopted, with eight electron emitters (two rows of four emitters) per module, wherein tungsten targets were brazed onto copper anode blocks—one anode block per module. A specialized ceramic connector provided high voltage standoff capability and cooling oil flow to the anode. A matrix topology and low-noise electronic controls provided switching of the emitters. Results: Four modules (32 x-ray sources in two rows of 16) have been successfully integrated into a single vacuum vessel and operated on an inverse-geometry computed tomography system. Dispenser cathodes provided high beam current (>1000 mA) in pulse mode, and the electrostatic lenses focused the current beam to a small optical focal spot size (0.5 × 1.4 mm). Controlled emitter grid voltage allowed the beam current to be varied for each source, providing the ability to modulate beam current across the fan of the x-ray beam, denoted as a virtual bowtie filter. The custom designed controls achieved x-ray source switching in <1 μs. The cathode-grounded source was operated successfully up to 120 kV. Conclusions: A high-power, distributed x-ray source for inverse-geometry CT applications was successfully designed, fabricated, and operated. Future embodiments may increase the number of spots and utilize fast read out detectors to increase the x-ray flux magnitude further, while still staying within the stationary target inherent

  14. Comparison of Tc-99m HM-PAO SPECT brain scan and x-ray CT in the detection of brain metastases

    International Nuclear Information System (INIS)

    Abdel-Dayem, H.M.; Sadek, S.; Sahwell, A.; Kubasek, H.; El-Sayed, M.; Ziada, G.; Mobarak, L.; Al-Huda, F.; Omar, Y.T.

    1986-01-01

    Tc-99m HM-PAO imaging was compared with x-ray CT in 14 patients with known or suspected brain metastases. Both studies were done within 3 days of each other. Static and single photon emission CT (SPECT) images were acquired after intravenous injection of 13 mCi of Tc-99m HM-PAO. All 14 patients underwent static and SPECT Tc-99m HM-PAO imaging and x-ray CT. Studies were positive in 7, 12, and 10 patients, respectively, by static, SPECT, and x-ray CT imaging, and negative in 7, 2, and 2. The number of lesions identified was 0 (static imaging), 32 (SPECT), and 26(x-ray CT). There were no ''suspicious'' studies by any modality. This study indicates that Tc-99m HM-PAO SPECT cerebral blood flow imaging is more sensitive than x-ray CT for detecting brain metastases, that biplane imaging is not sensitive and SPECT is essential, and that for Tc-99m HM-PAO SPECT brain imaging to regain its importance with respect to x-ray CT, acquisition time must be 10 minutes or less and determination of percentage brain uptake of the injected dose, and of regional distribution, is necessary

  15. Geometric calibration of a stationary digital breast tomosynthesis system based on distributed carbon nanotube X-ray source arrays.

    Directory of Open Access Journals (Sweden)

    Changhui Jiang

    Full Text Available Stationary digital breast tomosynthesis (sDBT with distributed X-ray sources based on carbon nanotube (CNT field emission cathodes has been recently proposed as an approach that can prevent motion blur produced by traditional DBT systems. In this paper, we simulate a geometric calibration method based on a proposed multi-source CNT X-ray sDBT system. This method is a projection matrix-based approach with seven geometric parameters, all of which can be obtained from only one projection datum of the phantom. To our knowledge, this study reports the first application of this approach in a CNT-based multi-beam X-ray sDBT system. The simulation results showed that the extracted geometric parameters from the calculated projection matrix are extremely close to the input values and that the proposed method is effective and reliable for a square sDBT system. In addition, a traditional cone-beam computed tomography (CT system was also simulated, and the uncalibrated and calibrated geometric parameters were used in image reconstruction based on the filtered back-projection (FBP method. The results indicated that the images reconstructed with calibrated geometric parameters have fewer artifacts and are closer to the reference image. All the simulation tests showed that this geometric calibration method is optimized for sDBT systems but can also be applied to other application-specific CT imaging systems.

  16. Geometric calibration of a stationary digital breast tomosynthesis system based on distributed carbon nanotube X-ray source arrays.

    Science.gov (United States)

    Jiang, Changhui; Zhang, Na; Gao, Juan; Hu, Zhanli

    2017-01-01

    Stationary digital breast tomosynthesis (sDBT) with distributed X-ray sources based on carbon nanotube (CNT) field emission cathodes has been recently proposed as an approach that can prevent motion blur produced by traditional DBT systems. In this paper, we simulate a geometric calibration method based on a proposed multi-source CNT X-ray sDBT system. This method is a projection matrix-based approach with seven geometric parameters, all of which can be obtained from only one projection datum of the phantom. To our knowledge, this study reports the first application of this approach in a CNT-based multi-beam X-ray sDBT system. The simulation results showed that the extracted geometric parameters from the calculated projection matrix are extremely close to the input values and that the proposed method is effective and reliable for a square sDBT system. In addition, a traditional cone-beam computed tomography (CT) system was also simulated, and the uncalibrated and calibrated geometric parameters were used in image reconstruction based on the filtered back-projection (FBP) method. The results indicated that the images reconstructed with calibrated geometric parameters have fewer artifacts and are closer to the reference image. All the simulation tests showed that this geometric calibration method is optimized for sDBT systems but can also be applied to other application-specific CT imaging systems.

  17. A preliminary study of breast cancer diagnosis using laboratory based small angle x-ray scattering

    Science.gov (United States)

    Round, A. R.; Wilkinson, S. J.; Hall, C. J.; Rogers, K. D.; Glatter, O.; Wess, T.; Ellis, I. O.

    2005-09-01

    Breast tissue collected from tumour samples and normal tissue from bi-lateral mastectomy procedures were examined using small angle x-ray scattering. Previous work has indicated that breast tissue disease diagnosis could be performed using small angle x-ray scattering (SAXS) from a synchrotron radiation source. The technique would be more useful to health services if it could be made to work using a conventional x-ray source. Consistent and reliable differences in x-ray scatter distributions were observed between samples from normal and tumour tissue samples using the laboratory based 'SAXSess' system. Albeit from a small number of samples, a sensitivity of 100% was obtained. This result encourages us to pursue the implementation of SAXS as a laboratory based diagnosis technique.

  18. A preliminary study of breast cancer diagnosis using laboratory based small angle x-ray scattering

    Energy Technology Data Exchange (ETDEWEB)

    Round, A R [Daresbury Laboratories, Warrington, WA4 4AD (United Kingdom); Wilkinson, S J [Daresbury Laboratories, Warrington, WA4 4AD (United Kingdom); Hall, C J [Daresbury Laboratories, Warrington, WA4 4AD (United Kingdom); Rogers, K D [Department of Materials and Medical Sciences, Cranfield University, Swindon, SN6 8LA (United Kingdom); Glatter, O [Department of Chemistry, University of Graz (Austria); Wess, T [School of Optometry and Vision Sciences, Cardiff University, Cardiff CF10 3NB, Wales (United Kingdom); Ellis, I O [Nottingham City Hospital, Nottingham (United Kingdom)

    2005-09-07

    Breast tissue collected from tumour samples and normal tissue from bi-lateral mastectomy procedures were examined using small angle x-ray scattering. Previous work has indicated that breast tissue disease diagnosis could be performed using small angle x-ray scattering (SAXS) from a synchrotron radiation source. The technique would be more useful to health services if it could be made to work using a conventional x-ray source. Consistent and reliable differences in x-ray scatter distributions were observed between samples from normal and tumour tissue samples using the laboratory based 'SAXSess' system. Albeit from a small number of samples, a sensitivity of 100% was obtained. This result encourages us to pursue the implementation of SAXS as a laboratory based diagnosis technique.

  19. A preliminary study of breast cancer diagnosis using laboratory based small angle x-ray scattering

    International Nuclear Information System (INIS)

    Round, A R; Wilkinson, S J; Hall, C J; Rogers, K D; Glatter, O; Wess, T; Ellis, I O

    2005-01-01

    Breast tissue collected from tumour samples and normal tissue from bi-lateral mastectomy procedures were examined using small angle x-ray scattering. Previous work has indicated that breast tissue disease diagnosis could be performed using small angle x-ray scattering (SAXS) from a synchrotron radiation source. The technique would be more useful to health services if it could be made to work using a conventional x-ray source. Consistent and reliable differences in x-ray scatter distributions were observed between samples from normal and tumour tissue samples using the laboratory based 'SAXSess' system. Albeit from a small number of samples, a sensitivity of 100% was obtained. This result encourages us to pursue the implementation of SAXS as a laboratory based diagnosis technique

  20. Low-Dose X-ray CT Reconstruction via Dictionary Learning

    Science.gov (United States)

    Xu, Qiong; Zhang, Lei; Hsieh, Jiang; Wang, Ge

    2013-01-01

    Although diagnostic medical imaging provides enormous benefits in the early detection and accuracy diagnosis of various diseases, there are growing concerns on the potential side effect of radiation induced genetic, cancerous and other diseases. How to reduce radiation dose while maintaining the diagnostic performance is a major challenge in the computed tomography (CT) field. Inspired by the compressive sensing theory, the sparse constraint in terms of total variation (TV) minimization has already led to promising results for low-dose CT reconstruction. Compared to the discrete gradient transform used in the TV method, dictionary learning is proven to be an effective way for sparse representation. On the other hand, it is important to consider the statistical property of projection data in the low-dose CT case. Recently, we have developed a dictionary learning based approach for low-dose X-ray CT. In this paper, we present this method in detail and evaluate it in experiments. In our method, the sparse constraint in terms of a redundant dictionary is incorporated into an objective function in a statistical iterative reconstruction framework. The dictionary can be either predetermined before an image reconstruction task or adaptively defined during the reconstruction process. An alternating minimization scheme is developed to minimize the objective function. Our approach is evaluated with low-dose X-ray projections collected in animal and human CT studies, and the improvement associated with dictionary learning is quantified relative to filtered backprojection and TV-based reconstructions. The results show that the proposed approach might produce better images with lower noise and more detailed structural features in our selected cases. However, there is no proof that this is true for all kinds of structures. PMID:22542666

  1. Comparative analysis among X-ray mammographic findings, nuclear and histologic grading, and TNM staging of breast carcinoma

    International Nuclear Information System (INIS)

    Park, Jin Sook; Sung, Ki Joon; Cho, Mee Yon; Hong, In Soo; Kim, Myung Soon; Oh, Ki Keun

    1996-01-01

    The purpose of this study was to evaluate the prognosis of breast carcinoma by comparison with X-ray mammographic findings, nuclear and histologic grade, and TNM staging. We retrospectively reviewed 114 cases (113 patients) of breast carcinoma, analysing X-ray mammographic findings of all cases with regard to mass, calcification, and spiculation. In 80 cases of scirrhous invasive ductal breast carcinoma. Black's nuclear and Bloom-Richardson's histologic grade were also evaluated. Mammographic findings and nuclear and histologic grade were compared with TNM staging which might suggest the prognosis of breast carcinoma. X-ray mammographic findings (mass, calcification and spiculation) did not significantly correlate with T staging, but the clinical staging of the spiculation was advanced. These X-ray findings did not significantly correlate with the nuclear grading and the histologic grading. Nuclear grade did not correlate with T and M staging, but correlated significantly with N staging and clinical stage(p < 0.05). Histologic grade did not significantly correlate with TNM staging. The clinical staging of spiculation was advanced and nuclear grade correlated significantly with N stage and clinical staging. X-ray mammographic findings did not directly correlate with nuclear and histologic grading, but combined studies of the evaluation of mammographic findings and nuclear and histologic grade were useful for prognosing breast carcinoma

  2. Pin-photodiode array for the measurement of fan-beam energy and air kerma distributions of X-ray CT scanners.

    Science.gov (United States)

    Haba, Tomonobu; Koyama, Shuji; Aoyama, Takahiko; Kinomura, Yutaka; Ida, Yoshihiro; Kobayashi, Masanao; Kameyama, Hiroshi; Tsutsumi, Yoshinori

    2016-07-01

    Patient dose estimation in X-ray computed tomography (CT) is generally performed by Monte Carlo simulation of photon interactions within anthropomorphic or cylindrical phantoms. An accurate Monte Carlo simulation requires an understanding of the effects of the bow-tie filter equipped in a CT scanner, i.e. the change of X-ray energy and air kerma along the fan-beam arc of the CT scanner. To measure the effective energy and air kerma distributions, we devised a pin-photodiode array utilizing eight channels of X-ray sensors arranged at regular intervals along the fan-beam arc of the CT scanner. Each X-ray sensor consisted of two plate type of pin silicon photodiodes in tandem - front and rear photodiodes - and of a lead collimator, which only allowed X-rays to impinge vertically to the silicon surface of the photodiodes. The effective energy of the X-rays was calculated from the ratio of the output voltages of the photodiodes and the dose was calculated from the output voltage of the front photodiode using the energy and dose calibration curves respectively. The pin-photodiode array allowed the calculation of X-ray effective energies and relative doses, at eight points simultaneously along the fan-beam arc of a CT scanner during a single rotation of the scanner. The fan-beam energy and air kerma distributions of CT scanners can be effectively measured using this pin-photodiode array. Copyright © 2016 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  3. The CT Scanner Facility at Stellenbosch University: An open access X-ray computed tomography laboratory

    Science.gov (United States)

    du Plessis, Anton; le Roux, Stephan Gerhard; Guelpa, Anina

    2016-10-01

    The Stellenbosch University CT Scanner Facility is an open access laboratory providing non-destructive X-ray computed tomography (CT) and a high performance image analysis services as part of the Central Analytical Facilities (CAF) of the university. Based in Stellenbosch, South Africa, this facility offers open access to the general user community, including local researchers, companies and also remote users (both local and international, via sample shipment and data transfer). The laboratory hosts two CT instruments, i.e. a micro-CT system, as well as a nano-CT system. A workstation-based Image Analysis Centre is equipped with numerous computers with data analysis software packages, which are to the disposal of the facility users, along with expert supervision, if required. All research disciplines are accommodated at the X-ray CT laboratory, provided that non-destructive analysis will be beneficial. During its first four years, the facility has accommodated more than 400 unique users (33 in 2012; 86 in 2013; 154 in 2014; 140 in 2015; 75 in first half of 2016), with diverse industrial and research applications using X-ray CT as means. This paper summarises the existence of the laboratory's first four years by way of selected examples, both from published and unpublished projects. In the process a detailed description of the capabilities and facilities available to users is presented.

  4. The CT Scanner Facility at Stellenbosch University: An open access X-ray computed tomography laboratory

    Energy Technology Data Exchange (ETDEWEB)

    Plessis, Anton du, E-mail: anton2@sun.ac.za [CT Scanner Facility, Central Analytical Facilities, Stellenbosch University, Stellenbosch (South Africa); Physics Department, Stellenbosch University, Stellenbosch (South Africa); Roux, Stephan Gerhard le, E-mail: lerouxsg@sun.ac.za [CT Scanner Facility, Central Analytical Facilities, Stellenbosch University, Stellenbosch (South Africa); Guelpa, Anina, E-mail: aninag@sun.ac.za [CT Scanner Facility, Central Analytical Facilities, Stellenbosch University, Stellenbosch (South Africa)

    2016-10-01

    The Stellenbosch University CT Scanner Facility is an open access laboratory providing non-destructive X-ray computed tomography (CT) and a high performance image analysis services as part of the Central Analytical Facilities (CAF) of the university. Based in Stellenbosch, South Africa, this facility offers open access to the general user community, including local researchers, companies and also remote users (both local and international, via sample shipment and data transfer). The laboratory hosts two CT instruments, i.e. a micro-CT system, as well as a nano-CT system. A workstation-based Image Analysis Centre is equipped with numerous computers with data analysis software packages, which are to the disposal of the facility users, along with expert supervision, if required. All research disciplines are accommodated at the X-ray CT laboratory, provided that non-destructive analysis will be beneficial. During its first four years, the facility has accommodated more than 400 unique users (33 in 2012; 86 in 2013; 154 in 2014; 140 in 2015; 75 in first half of 2016), with diverse industrial and research applications using X-ray CT as means. This paper summarises the existence of the laboratory’s first four years by way of selected examples, both from published and unpublished projects. In the process a detailed description of the capabilities and facilities available to users is presented.

  5. TU-CD-207-11: Patient-Driven Automatic Exposure Control for Dedicated Breast CT

    International Nuclear Information System (INIS)

    Hernandez, A; Gazi, P; Seibert, J; Boone, J

    2015-01-01

    Purpose: To implement automatic exposure control (AEC) in dedicated breast CT (bCT) on a patient-specific basis using only the pre-scan scout views. Methods: Using a large cohort (N=153) of bCT data sets, the breast effective diameter (D) and width in orthogonal planes (Wa,Wb) were calculated from the reconstructed bCT image and pre-scan scout views, respectively. D, Wa, and Wb were measured at the breast center-of-mass (COM), making use of the known geometry of our bCT system. These data were then fit to a second-order polynomial “D=F(Wa,Wb)” in a least squares sense in order to provide a functional form for determining the breast diameter. The coefficient of determination (R 2 ) and mean percent error between the measured breast diameter and fit breast diameter were used to evaluate the overall robustness of the polynomial fit. Lastly, previously-reported bCT technique factors derived from Monte Carlo simulations were used to determine the tube current required for each breast diameter in order to match two-view mammographic dose levels. Results: F(Wa,Wb) provided fitted breast diameters in agreement with the measured breast diameters resulting in R 2 values ranging from 0.908 to 0.929 and mean percent errors ranging from 3.2% to 3.7%. For all 153 bCT data sets used in this study, the fitted breast diameters ranged from 7.9 cm to 15.7 cm corresponding to tube current values ranging from 0.6 mA to 4.9 mA in order to deliver the same dose as two-view mammography in a 50% glandular breast with a 80 kV x-ray beam and 16.6 second scan time. Conclusion: The present work provides a robust framework for AEC in dedicated bCT using only the width measurements derived from the two orthogonal pre-scan scout views. Future work will investigate how these automatically chosen exposure levels affect the quality of the reconstructed image

  6. TU-CD-207-11: Patient-Driven Automatic Exposure Control for Dedicated Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Hernandez, A; Gazi, P [Biomedical Engineering Graduate Group, University of California Davis, Davis, CA (United States); Department of Radiology, UC Davis Medical Center, Sacramento, CA (United States); Seibert, J; Boone, J [Department of Radiology, UC Davis Medical Center, Sacramento, CA (United States); Department of Biomedical Engineering, University of California Davis, Davis, CA (United States)

    2015-06-15

    Purpose: To implement automatic exposure control (AEC) in dedicated breast CT (bCT) on a patient-specific basis using only the pre-scan scout views. Methods: Using a large cohort (N=153) of bCT data sets, the breast effective diameter (D) and width in orthogonal planes (Wa,Wb) were calculated from the reconstructed bCT image and pre-scan scout views, respectively. D, Wa, and Wb were measured at the breast center-of-mass (COM), making use of the known geometry of our bCT system. These data were then fit to a second-order polynomial “D=F(Wa,Wb)” in a least squares sense in order to provide a functional form for determining the breast diameter. The coefficient of determination (R{sup 2}) and mean percent error between the measured breast diameter and fit breast diameter were used to evaluate the overall robustness of the polynomial fit. Lastly, previously-reported bCT technique factors derived from Monte Carlo simulations were used to determine the tube current required for each breast diameter in order to match two-view mammographic dose levels. Results: F(Wa,Wb) provided fitted breast diameters in agreement with the measured breast diameters resulting in R{sup 2} values ranging from 0.908 to 0.929 and mean percent errors ranging from 3.2% to 3.7%. For all 153 bCT data sets used in this study, the fitted breast diameters ranged from 7.9 cm to 15.7 cm corresponding to tube current values ranging from 0.6 mA to 4.9 mA in order to deliver the same dose as two-view mammography in a 50% glandular breast with a 80 kV x-ray beam and 16.6 second scan time. Conclusion: The present work provides a robust framework for AEC in dedicated bCT using only the width measurements derived from the two orthogonal pre-scan scout views. Future work will investigate how these automatically chosen exposure levels affect the quality of the reconstructed image.

  7. Quasimonochromatic x-ray computed tomography by the balanced filter method using a conventional x-ray source

    International Nuclear Information System (INIS)

    Saito, Masatoshi

    2004-01-01

    A quasimonochromatic x-ray computed tomography (CT) system utilizing balanced filters has recently been developed for acquiring quantitative CT images. This system consisted of basic components such as a conventional x-ray generator for radiography, a stage for mounting and rotating objects, and an x-ray line sensor camera. Metallic sheets of Er and Yb were used as the balanced filters for obtaining quasimonochromatic incident x rays that include the characteristic lines of the W Kα doublet from a tungsten target. The mean energy and energy width of the quasimonochromatic x rays were determined to be 59.0 and 1.9 keV, respectively, from x-ray spectroscopic measurements using a high-purity Ge detector. The usefulness of the present x-ray CT system was demonstrated by obtaining spatial distributions of the linear attenuation coefficients of three selected samples--a 20 cm diameter cylindrical water phantom, a 3.5 cm diameter aluminum rod, and a human head phantom. The results clearly indicate that this apparatus is surprisingly effective for estimating the distribution of the linear attenuation coefficients without any correction of the beam-hardening effect. Thus, implementing the balanced filter method on an x-ray CT scanner has promise in producing highly quantitative CT images

  8. Transference of surface markers in X-ray, CT- and MR-imaging

    International Nuclear Information System (INIS)

    Grevelhoerster, T.; Poetter, R.; Prott, F.J.; Dittrich, M.

    1995-01-01

    By using pharmacological capsules and plastic tubes filled with oily contrast medium contraining iodine (Lipiodol; Byk Gulden), marking aids were developed which can be seen in similar definite limits within the framework of MRI-, CT- and conventional X-ray-Imaging. A method to view these new, artificial markers in combination with individual, anatomical landmarks is introduced. The marking aid/surface marker, fixed on anatomical reference structures on the skin, does not result in an additional burden for the patient. The new, artificial markers are also useful for making other structures recognizable, such as anatomical relation lines, center of the portal and edges in planning imaging for radiotherapy treatments and are used as leading and reference structures to compare localisation and extent of lesions in X-ray-, CT- and MRI. Marking aids/surface markers do not have to be changed in different imaging methods. (orig./MG) [de

  9. X-ray micro-CT and neutron CT as complementary imaging tools for non-destructive 3D imaging of rare silicified fossil plants

    Science.gov (United States)

    Karch, J.; Dudák, J.; Žemlička, J.; Vavřík, D.; Kumpová, I.; Kvaček, J.; Heřmanová, Z.; Šoltés, J.; Viererbl, L.; Morgano, M.; Kaestner, A.; Trtík, P.

    2017-12-01

    Computed tomography provides 3D information of inner structures of investigated objects. The obtained information is, however, strongly dependent on the used radiation type. It is known that as X-rays interact with electron cloud and neutrons with atomic nucleus, the obtained data often provide different contrast of sample structures. In this work we present a set of comparative radiographic and CT measurements of rare fossil plant samples using X-rays and thermal neutrons. The X-ray measurements were performed using large area photon counting detectors Timepix at IEAP CTU in Prague and Perkin Elmer flat-panel detector at Center of Excellence Telč. The neutron CT measurement was carried out at Paul Scherrer Institute using BOA beam-line. Furthermore, neutron radiography of fossil samples, provided by National Museum, were performed using a large-area Timepix detector with a neutron-sensitive converting 6LiF layer at Research Centre Rez, Czech Republic. The obtained results show different capabilities of both imaging approaches. While X-ray micro-CT provides very high resolution and enables visualization of fine cracks or small cavities in the samples neutron imaging provides high contrast of morphological structures of fossil plant samples, where X-ray imaging provides insufficient contrast.

  10. Direct 3-D morphological measurements of silicone rubber impression using micro-focus X-ray CT.

    Science.gov (United States)

    Kamegawa, Masayuki; Nakamura, Masayuki; Fukui, Yu; Tsutsumi, Sadami; Hojo, Masaki

    2010-01-01

    Three-dimensional computer models of dental arches play a significant role in prosthetic dentistry. The microfocus X-ray CT scanner has the advantage of capturing precise 3D shapes of deep fossa, and we propose a new method of measuring the three-dimensional morphology of a dental impression directly, which will eliminate the conversion process to dental casts. Measurement precision and accuracy were evaluated using a standard gage comprised of steel balls which simulate the dental arch. Measurement accuracy, standard deviation of distance distribution of superimposed models, was determined as +/-0.050 mm in comparison with a CAD model. Impressions and casts of an actual dental arch were scanned by microfocus X-ray CT and three-dimensional models were compared. The impression model had finer morphology, especially around the cervical margins of teeth. Within the limitations of the current study, direct three-dimensional impression modeling was successfully demonstrated using microfocus X-ray CT.

  11. X-ray absorption microtomography (microCT) and small beam diffraction mapping of sea urchin teeth.

    Science.gov (United States)

    Stock, S R; Barss, J; Dahl, T; Veis, A; Almer, J D

    2002-07-01

    Two noninvasive X-ray techniques, laboratory X-ray absorption microtomography (microCT) and X-ray diffraction mapping, were used to study teeth of the sea urchin Lytechinus variegatus. MicroCT revealed low attenuation regions at near the tooth's stone part and along the carinar process-central prism boundary; this latter observation appears to be novel. The expected variation of Mg fraction x in the mineral phase (calcite, Ca(1-x)Mg(x)CO(3)) cannot account for all of the linear attenuation coefficient decrease in the two zones: this suggested that soft tissue is localized there. Transmission diffraction mapping (synchrotron X-radiation, 80.8 keV, 0.1 x 0.1mm(2) beam area, 0.1mm translation grid, image plate area detector) simultaneously probed variations in 3-D and showed that the crystal elements of the "T"-shaped tooth were very highly aligned. Diffraction patterns from the keel (adaxial web) and from the abaxial flange (containing primary plates and the stone part) differed markedly. The flange contained two populations of identically oriented crystal elements with lattice parameters corresponding to x=0.13 and x=0.32. The keel produced one set of diffraction spots corresponding to the lower x. The compositions were more or less equivalent to those determined by others for camarodont teeth, and the high Mg phase is expected to be disks of secondary mineral epitaxially related to the underlying primary mineral element. Lattice parameter gradients were not noted in the keel or flange. Taken together, the microCT and diffraction results indicated that there was a band of relatively high protein content, of up to approximately 0.25 volume fraction, in the central part of the flange and paralleling its adaxial and abaxial faces. X-ray microCT and microdiffraction data used in conjunction with protein distribution data will be crucial for understanding the properties of various biocomposites and their mechanical functions.

  12. SU-E-T-70: Commissioning a Multislice CT Scanner for X-Ray CT Polymer Gel Dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Johnston, H [University of Victoria, Victoria, BC (Australia); UT Southwestern Medical Center, Dallas, TX (United States); Hilts, M [University of Victoria, Victoria, BC (Australia); BC Cancer Agency, Kelowna, BC (Australia); Jirasek, A [University of Victoria, Victoria, BC (Australia)

    2014-06-01

    Purpose: To commission a multislice computed tomography (CT) scanner for fast and reliable readout of radiation therapy (RT) dose distributions using CT polymer gel dosimetry (PGD). Methods: Commissioning was performed for a 16-slice CT scanner using images acquired through a 1L cylinder filled with water. Additional images were collected using a single slice machine for comparison purposes. The variability in CT number associated with the anode heel effect was evaluated and used to define a new slice-by-slice background image subtraction technique. Image quality was assessed for the multislice system by comparing image noise and uniformity to that of the single slice machine. The consistency in CT number across slices acquired simultaneously using the multislice detector array was also evaluated. Finally, the variability in CT number due to increasing x-ray tube load was measured for the multislice scanner and compared to the tube load effects observed on the single slice machine. Results: Slice-by-slice background subtraction effectively removes the variability in CT number across images acquired simultaneously using the multislice scanner and is the recommended background subtraction method when using a multislice CT system. Image quality for the multislice machine was found to be comparable to that of the single slice scanner. Further study showed CT number was consistent across image slices acquired simultaneously using the multislice detector array for each detector configuration of the slice thickness examined. In addition, the multislice system was found to eliminate variations in CT number due to increasing x-ray tube load and reduce scanning time by a factor of 4 when compared to imaging a large volume using a single slice scanner. Conclusion: A multislice CT scanner has been commissioning for CT PGD, allowing images of an entire dose distribution to be acquired in a matter of minutes. Funding support provided by the Natural Sciences and Engineering

  13. Determination of x-ray spectra incident on and transmitted through whole excised breast for improved mammographic exposing technique

    International Nuclear Information System (INIS)

    Haus, A.G.; Metz, C.E.; Doi, K.; Bernstein, J.

    1976-01-01

    Measurements were made of x-ray spectra incident on and transmitted through whole excised fresh breast tissue. The spectra and corresponding radiographs show that hardening of the x-ray beam from a molybdenum anode x-ray tube by a 0.5 mm Al filter instead of the conventional 0.03 mm Mo filter allows shorter exposure times and reduces patient radiation exposure. With the Al filter, acceptable image contrast is preserved in mammography of thick or dense breasts

  14. WE-FG-207A-00: Advances in Dedicated Breast CT

    International Nuclear Information System (INIS)

    2016-01-01

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  15. WE-FG-207A-00: Advances in Dedicated Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2016-06-15

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  16. Diagnostic modalities x-ray and CT chest differ in the management of thoracic injury

    Directory of Open Access Journals (Sweden)

    D Chapagain

    2015-06-01

    Full Text Available Objective: To observe difference in the management of blunt trauma to the chest on the basis of conventional xray and computerised tomography of the chest. Methods: This prospective study was conducted between December 2011 to October 2012 in COMS in Bharatpur,a tertiary referral centre in central Nepal . Clinically stable thoracic injury patients were first evaluated with chest x-ray and the management on this basis was recorded. The findings of the CT chest were assessed and the type of management on the basis of CT was also recorded. Outcome was assessed in terms of mortality, morbidity, hospital and ICU stay with respect to the management on the basis of chest x-ray and CT scan. Results: Of the 129 patients, 74.4% were male and 25.6% were female with the patients ranging in age from 7 to 87 years (mean = 40.41 years. The most common mechanism of trauma to the chest was as a result of a motor vehicle accident (69.8%, followed by fall injury (20.2%. X-ray chest diagnosed rib fracture in 62%, haemothorax in 37%, pneumothorax in 27%, lung contusion in 10% and haemopneumothorax in 21% patients. Similarly CT chest diagnosed rib fracture in 86%, haemothorax in 54%, pneumothorax in 36%, lung contusion in 30% and haemopneumothorax in 30% patients. Mean hospital stay was 9.5 days in the group of patients having management on the basis of x-ray chest relative to mean stay of 10.2 days in the CT- chest group. In the management on the basis of xray group, there was a mean ICU stay of 2.8days compared to mean stays of 3.2 days in CT chest group. Conclusion: Though CT scan of the chest is more informative and differs the management of the blunt chest trauma, one should not forget to advise the cost effective, easily available and initial guiding agent, xray chest for early management of the chest injury patient. DOI: http://dx.doi.org/10.3126/jcmsn.v10i1.12764 Journal of College of Medical Sciences-Nepal, 2014, Vol.10(1; 22-31

  17. Incorporating multislice imaging into x-ray CT polymer gel dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Johnston, H., E-mail: holly.johnston@utsw.edu [Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8W 2Y2 (Canada); Hilts, M. [Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8W 2Y2, Canada and Medical Physics, BC Cancer Agency, Vancouver Island Centre, Victoria, British Columbia V8R 6V5 (Canada); Jirasek, A. [Department of Physics and Astronomy, University of Victoria, Victoria, British Columbia V8W 2Y2, Canada and Department of Physics, University of British Columbia—Okanagan Campus, Kelowna, British Columbia V1V 1V7 (Canada)

    2015-04-15

    Purpose: To evaluate multislice computed tomography (CT) scanning for fast and reliable readout of radiation therapy (RT) dose distributions using CT polymer gel dosimetry (PGD) and to establish a baseline assessment of image noise and uniformity in an unirradiated gel dosimeter. Methods: A 16-slice CT scanner was used to acquire images through a 1 L cylinder filled with water. Additional images were collected using a single slice machine. The variability in CT number (N{sub CT}) associated with the anode heel effect was evaluated and used to define a new slice-by-slice background subtraction artifact removal technique for CT PGD. Image quality was assessed for the multislice system by evaluating image noise and uniformity. The agreement in N{sub CT} for slices acquired simultaneously using the multislice detector array was also examined. Further study was performed to assess the effects of increasing x-ray tube load on the constancy of measured N{sub CT} and overall scan time. In all cases, results were compared to the single slice machine. Finally, images were collected throughout the volume of an unirradiated gel dosimeter to quantify image noise and uniformity before radiation is delivered. Results: Slice-by-slice background subtraction effectively removes the variability in N{sub CT} observed across images acquired simultaneously using the multislice scanner and is the recommended background subtraction method when using a multislice CT system. Image noise was higher for the multislice system compared to the single slice scanner, but overall image quality was comparable between the two systems. Further study showed N{sub CT} was consistent across image slices acquired simultaneously using the multislice detector array for each detector configuration of the slice thicknesses examined. In addition, the multislice system was found to eliminate variations in N{sub CT} due to increasing x-ray tube load and reduce scanning time by a factor of 4 when compared to

  18. Cone-beam volume CT mammographic imaging: feasibility study

    Science.gov (United States)

    Chen, Biao; Ning, Ruola

    2001-06-01

    X-ray projection mammography, using a film/screen combination or digital techniques, has proven to be the most effective imaging modality for early detection of breast cancer currently available. However, the inherent superimposition of structures makes small carcinoma (a few millimeters in size) difficult to detect in the occultation case or in dense breasts, resulting in a high false positive biopsy rate. The cone-beam x-ray projection based volume imaging using flat panel detectors (FPDs) makes it possible to obtain three-dimensional breast images. This may benefit diagnosis of the structure and pattern of the lesion while eliminating hard compression of the breast. This paper presents a novel cone-beam volume CT mammographic imaging protocol based on the above techniques. Through computer simulation, the key issues of the system and imaging techniques, including the x-ray imaging geometry and corresponding reconstruction algorithms, x-ray characteristics of breast tissues, x-ray setting techniques, the absorbed dose estimation and the quantitative effect of x-ray scattering on image quality, are addressed. The preliminary simulation results support the proposed cone-beam volume CT mammographic imaging modality in respect to feasibility and practicability for mammography. The absorbed dose level is comparable to that of current two-view mammography and would not be a prominent problem for this imaging protocol. Compared to traditional mammography, the proposed imaging protocol with isotropic spatial resolution will potentially provide significantly better low contrast detectability of breast tumors and more accurate location of breast lesions.

  19. Design of scanning motion control system for high-energy X-ray industrial CT

    International Nuclear Information System (INIS)

    Duan Liming

    2008-01-01

    A scanning motion control system was developed for the high-energy X-ray industrial computerized tomography (CT). The system consists of an industrial control computer, a counter card, a control card, servo drivers, servo motors, working platforms, gratings and control software. Based on windows driver model(WDM) mode, the composition of the driver pro- gram for the system was studied. Took the motor control card as an example, the method to develop the driver program was researched, and the intercourse process between the device driver program and the user-program was analyzed. The real-time control of the system was implemented using the WDM driver. The real-time performance and reliability of the system can satisfy the requirement of high-energy X-ray industrial CT. (authors)

  20. Measurement of position dependence of spectral distribution in primary X-ray beam of CT system using compton spectroscopy

    International Nuclear Information System (INIS)

    Matsumoto, Masao; Maeda, Koji; Maeda, Koji

    2005-01-01

    Our purpose is to acquire the diagnostic x-ray spectra of the CT system easily under clinical conditions by Compton scatter spectroscopy using a high resolution Schottky CdTe detector recently developed and to estimate the quality and quantity of these spectra compared the relative exposure measured using an ionization chamber with curves measured by the reconstructed spectra. Although the spectral broadening and the fluctuation were slightly shown, the errors of the relative exposure were under 10%. Though the experimental arrangement of this method is so easy and reconstructed algorithm is simple, these results suggested that this method is little influenced by the variation of dose rates or property of the incident x rays. We therefore propose that the Compton scatter spectroscopy with the newly developed Schottky CdTe detector is suitable for measuring the diagnostic x-ray spectra of the CT system under the clinical conditions and useful for quality assurance and quality control of the clinical x-ray CT system. (author)

  1. X-ray phase-contrast CT of a pancreatic ductal adenocarcinoma mouse model.

    Directory of Open Access Journals (Sweden)

    Arne Tapfer

    Full Text Available To explore the potential of grating-based x-ray phase-contrast computed tomography (CT for preclinical research, a genetically engineered mouse model of pancreatic ductal adenocarcinoma (PDAC was investigated. One ex-vivo mouse specimen was scanned with different grating-based phase-contrast CT imaging setups covering two different settings: i high-resolution synchrotron radiation (SR imaging and ii dose-reduced imaging using either synchrotron radiation or a conventional x-ray tube source. These experimental settings were chosen to assess the potential of phase-contrast imaging for two different types of application: i high-performance imaging for virtual microscopy applications and ii biomedical imaging with increased soft-tissue contrast for in-vivo applications. For validation and as a reference, histological slicing and magnetic resonance imaging (MRI were performed on the same mouse specimen. For each x-ray imaging setup, attenuation and phase-contrast images were compared visually with regard to contrast in general, and specifically concerning the recognizability of lesions and cancerous tissue. To quantitatively assess contrast, the contrast-to-noise ratios (CNR of selected regions of interest (ROI in the attenuation images and the phase images were analyzed and compared. It was found that both for virtual microscopy and for in-vivo applications, there is great potential for phase-contrast imaging: in the SR-based benchmarking data, fine details about tissue composition are accessible in the phase images and the visibility of solid tumor tissue under dose-reduced conditions is markedly superior in the phase images. The present study hence demonstrates improved diagnostic value with phase-contrast CT in a mouse model of a complex endogenous cancer, promoting the use and further development of grating-based phase-contrast CT for biomedical imaging applications.

  2. TU-G-207-00: Emerging Applications of X-Ray Imaging

    International Nuclear Information System (INIS)

    2015-01-01

    Last few years has witnessed the development of novel of X-ray imaging modalities, such as spectral CT, phase contrast CT, and X-ray acoustic/fluorescence/luminescence imaging. This symposium will present the recent advances of these emerging X-ray imaging modalities and update the attendees with knowledge in various related topics, including X-ray photon-counting detectors, X-ray physics underlying the emerging applications beyond the traditional X-ray imaging, image reconstruction for the novel modalities, characterization and evaluation of the systems, and their practical implications. In addition, the concept and practical aspects of X-ray activatable targeted nanoparticles for molecular X-ray imaging will be discussed in the context of X-ray fluorescence and luminescence CT. Learning Objectives: Present background knowledge of various emerging X-ray imaging techniques, such as spectral CT, phase contrast CT and X-ray fluorescence/luminescence CT. Discuss the practical need, technical aspects and current status of the emerging X-ray imaging modalities. Describe utility and future impact of the new generation of X-ray imaging applications

  3. Synchrotron X-ray CT of rose peduncles. Evaluation of tissue damage by radiation

    International Nuclear Information System (INIS)

    Herppich, Werner B.; Zabler, Simon; Dawson, Martin; Choinka, Gerard; Manke, Ingo

    2015-01-01

    ''Bent-neck'' syndrome, an important postharvest problem of cut roses, is probably caused by water supply limitations and/or the structural weakness of vascular bundles of the peduncle tissue. For this reason, advanced knowledge about the microstructures of rose peduncles and their cultivar specific variations may lead to a better understanding of the underlying mechanisms. Synchrotron X-ray computed tomography (SXCT), especially phase-based CT, is a highly suitable technique to nondestructively investigate plants' micro anatomy. SXCT with monochromatic X-ray beams of 30, 40 and 50 keV photon energy was used to evaluate the three-dimensional inner structures of the peduncles of 3 rose cultivars that differ greatly in their bent-neck susceptibility. Results indicated that this technique achieves sufficiently high spatial resolution to investigate complex tissues. However, further investigations with chlorophyll fluorescence analysis (CFA) and optical microscope imagery reveal different kinds of heavy damage of the irradiated regions induced by synchrotron X-rays; in a cultivar-specific manner, partial destruction of cell walls occurred a few hours after X-ray irradiation. Furthermore, a delayed inhibition of photosynthesis accompanied by the degradation of chlorophyll was obvious from CFA within hours and days after the end of CT measurements. Although SXCT is certainly well suited for three-dimensional anatomical analysis of rose peduncles, the applied technique is not nondestructive.

  4. Synchrotron X-ray CT of rose peduncles. Evaluation of tissue damage by radiation

    Energy Technology Data Exchange (ETDEWEB)

    Herppich, Werner B. [Leibniz-Institut fuer Agrartechnik Potsdam-Bornim e.V., Potsdam (Germany). Abt. Technik im Gartenbau; Matsushima, Uzuki [Iwate Univ., Morioka (Japan). Faculty of Agriculture; Graf, Wolfgang [Association for Technology and Structures in Agriculture (KTBL), Darmstadt (Germany); Zabler, Simon [Fraunhofer-Institut fuer Integrierte Schaltungen (IIS), Wuerzburg (Germany). Project group NanoCT Systems (NCTS); Dawson, Martin [Salford Univ., Greater Manchester (United Kingdom); Choinka, Gerard; Manke, Ingo [Helmholtz Center Berlin for Materials and Energy (HZB), Berlin (Germany)

    2015-02-01

    ''Bent-neck'' syndrome, an important postharvest problem of cut roses, is probably caused by water supply limitations and/or the structural weakness of vascular bundles of the peduncle tissue. For this reason, advanced knowledge about the microstructures of rose peduncles and their cultivar specific variations may lead to a better understanding of the underlying mechanisms. Synchrotron X-ray computed tomography (SXCT), especially phase-based CT, is a highly suitable technique to nondestructively investigate plants' micro anatomy. SXCT with monochromatic X-ray beams of 30, 40 and 50 keV photon energy was used to evaluate the three-dimensional inner structures of the peduncles of 3 rose cultivars that differ greatly in their bent-neck susceptibility. Results indicated that this technique achieves sufficiently high spatial resolution to investigate complex tissues. However, further investigations with chlorophyll fluorescence analysis (CFA) and optical microscope imagery reveal different kinds of heavy damage of the irradiated regions induced by synchrotron X-rays; in a cultivar-specific manner, partial destruction of cell walls occurred a few hours after X-ray irradiation. Furthermore, a delayed inhibition of photosynthesis accompanied by the degradation of chlorophyll was obvious from CFA within hours and days after the end of CT measurements. Although SXCT is certainly well suited for three-dimensional anatomical analysis of rose peduncles, the applied technique is not nondestructive.

  5. X- Ray marking of breast cancer in Mongolia

    International Nuclear Information System (INIS)

    Ranish, A.; Tugsjargal, P.; Tuvshinjargal, D.; Nyamdavaa, N.; Gonchigsuren, D.

    2007-01-01

    Full text: Mammography and other breast imaging techniques such as ultrasound are useful in the evaluation of women who have a sign or symptom that may suggest breast cancer. Physical examination evaluates different tissue characteristics than mammography. We have evaluated mammograms of 1286 patients with breast diseases and detected 107 patients with breast cancer. 25 patients underwent surgery and approved all of mammography diagnosis. In all 122 patients with diagnosed breast cancer, the normal breast X-ray markings have been changed; in 107 cases there were single nodular foci, in 15 patients infiltrative cancer was observed. In 9 cases with absence of cancer lesion on mammography, its secondary signs diagnosed the breast cancer. In 50,8% (62) of cases the visible lesion of the breast cancer was star-shaped, 25.4% (31)- lobed, 16,4% (20)- ovoid, 7,4% (9)- round. The lesion contrast varied differently due to its size; in 58,1%- the lesion was intensive, in 11%- poorly intensive, in 4.7% of all cases the cancer lesion was very vague. In mammography, we observed sprinkled-salt like calcification (7,3%), linear calcification (3,2%) and also irregular calcifications (1,6%). Mammography showed its effectiveness in the early diagnosis of breast cancer. In many cases, the radiographic differentiation between benign and malignant circumscribed masses was difficult or sometimes impossible and we have used other diagnostic modalities like mammography guided biopsy. In many cases, the radiographic differentiation between benign and malignant circumscribed masses was difficult or sometimes impossible and we have used other diagnostic modalities like mammography guided biopsy. (author)

  6. Optimizing abdominal CT dose and image quality with respect to x-ray tube voltage

    Science.gov (United States)

    Huda, Walter; Ogden, Kent M.

    2004-05-01

    The objective of this study was to identify the x-ray tube voltage that results in optimum performance for abdominal CT imaging for a range of imaging tasks and patient sizes. Theoretical calculations were performed of the contrast to noise ratio (CNR) for disk shaped lesions of muscle, fat, bone and iodine embedded in a uniform water background. Lesion contrast was the mean Hounsfield Unit value at the effective photon energy, and image noise was determined from the total radiation intensity incident on the CT x-ray detector. Patient size ranging from young infants (10 kg) to oversized adults (120 kg), with CNR values obtained for x-ray tube voltages ranging from 80 to 140 kV. Patients of varying sizes were modeled as an equivalent cylinder of water, and the mean section dose (D) was determined for each selected x-ray tube kV value at a constant mAs. For each patient size and lesion type, we identified an optimal kV as the x-ray tube voltage that yields a maximum value of the figure of merit (CNR2/D). Increasing the x-ray tube voltage from 80 to 140 kV reduced lesion contrast by 11% for muscle, 21% for fat, 35% for bone and 52% for iodine, and these reductions were approximately independent of patient size. Increasing the x-ray tube voltage from 80 to 140 kV increased a muscle lesion CNR relative to a uniform water background by a factor of 2.6, with similar trends observed for fat (2.3), bone (1.9) and iodine (1.4). The improvement in lesion CNR with increasing x-ray tube voltage was highest for the largest sized patients. Increasing the x-ray tube voltage from 80 to 140 kV increased the patient dose by a factor of between 5.0 and 6.2 depending on the patient size. For small sized patients (10 and 30 kg) and muscle lesions, best performance is obtained at 80 kV; however, for adults (70 kg) and oversized adults (120 kg), the best performance would be obtained at 140 kV. Imaging fat lesions was best performed at 80 kV for all patients except for oversized adults

  7. Assessing the registration of CT-scan data to intraoperative x rays by fusing x rays and preoperative information

    Science.gov (United States)

    Gueziec, Andre P.

    1999-05-01

    This paper addresses a key issue of providing clinicians with visual feedback to validate a computer-generated registration of pre-operative and intra-operative data. With this feedback information, the clinician may decide to proceed with a computer-assisted intervention, revert to a manual intervention, or potentially provide information to the computer system to improve the registration. The paper focuses on total hip replacement (THR) surgery, but similar techniques could be applied to other types of interventions or therapy, including orthopedics, neurosurgery, and radiation therapy. Pre-operative CT data is used to plane the surgery (select an implant type, size and precise position), and is registered to intra-operative X-ray images, allowing to execute the plan: mill a cavity with the implant's shape. (Intra-operative X-ray images must be calibrated with respect to the surgical device executing the plan). One novel technique presented in this paper consists of simulating a post-operative X-ray image of the tissue of interest before doing the procedure, by projecting the registered implant onto an intra-operative X- ray image (corrected for distortion or not), providing clinicians with familiar and easy to interpret images. As an additional benefit, this method provides new means for comparing various strategies for registering pre-operative data to the physical space of the operating room.

  8. Development of high-resolution x-ray CT system using parallel beam geometry

    Energy Technology Data Exchange (ETDEWEB)

    Yoneyama, Akio, E-mail: akio.yoneyama.bu@hitachi.com; Baba, Rika [Central Research Laboratory, Hitachi Ltd., Hatoyama, Saitama (Japan); Hyodo, Kazuyuki [Institute of Materials Science, High Energy Accelerator Research Organization, Tsukuba, Ibaraki (Japan); Takeda, Tohoru [School of Allied Health Sciences, Kitasato University, Sagamihara, Kanagawa (Japan); Nakano, Haruhisa; Maki, Koutaro [Department of Orthodontics, School of Dentistry Showa University, Ota-ku, Tokyo (Japan); Sumitani, Kazushi; Hirai, Yasuharu [Kyushu Synchrotron Light Research Center, Tosu, Saga (Japan)

    2016-01-28

    For fine three-dimensional observations of large biomedical and organic material samples, we developed a high-resolution X-ray CT system. The system consists of a sample positioner, a 5-μm scintillator, microscopy lenses, and a water-cooled sCMOS detector. Parallel beam geometry was adopted to attain a field of view of a few mm square. A fine three-dimensional image of birch branch was obtained using a 9-keV X-ray at BL16XU of SPring-8 in Japan. The spatial resolution estimated from the line profile of a sectional image was about 3 μm.

  9. Status of X-ray CT photography in the radiology department of our school of dentistry

    International Nuclear Information System (INIS)

    Hokari, Seishi; Takahashi, Nobutoshi; Tsutsumi, Hiroyuki; Shirai, Yoshihiro; Inoue, Hiroshi; Inoue, Nobuyuki; Yamada, Hidehiko; Okumura, Yasuhiko

    1999-01-01

    Aided by the significant increase in its performance and reduction in cost, the X-ray CT, which originated from the EMI scanner of the 1970's, has become an essential tool in the field of diagnostic X-ray images in medicine. To comprehend in detail the status of X-ray CT photography used in a special diagnostic situation of an adjust hospital of a school of dentistry is beneficial for evaluating the use of exploratory X-ray photography to obtain useful diagnostic information as well as routine photography format. Also, knowledge of its status is necessary to address the issues of justification and optimization of X-ray examinations. Therefore, we conducted research and evaluation on various items, such as the number of cases photographed, number of contrast enhanced, and male to female ratio, over a six-years period to assess the status of photography conducted with a Toshiba CT Scanner TCT-700S, which was introduced to the Radiology Department of our school in September 1991. The following are part of our findings; 1. Status of photography in total. 1) The total number of cases photographed was 5,346 cases. 2) The number of cases contrast enhanced was 903 cases. 3) Contrast enhanced cases ratio was 16.9%. 4) Male to female ratio was 55.4: 44.6%. 2. Status of case photography by different department was: dentistry, 73.3%; medicine, 26.7%. In order to descending case frequency was oral surgery, 53.0%; internal medicine, 19.4%; and radiology 15.0%. 3. Status of case photography by age was, in order of descending frequency, 50s, 40s, and 60s. (author)

  10. Dual-energy X-ray micro-CT imaging of hybrid Ni/Al open-cell foam

    Czech Academy of Sciences Publication Activity Database

    Fíla, T.; Kumpová, Ivana; Koudelka_ml., P.; Zlámal, P.; Vavřík, Daniel; Jiroušek, O.; Jung, A.

    2016-01-01

    Roč. 11, č. 1 (2016), C01005 ISSN 1748-0221. [International Workshop on Radiation Imaging Detectors (IWORID2015) /17./. Hamburg, 28.06.2015-02.07.2015] Institutional support: RVO:68378297 Keywords : computerized tomography (CT) * computed radiography (CR) * X-ray radiography and digital radiography (DR) * inspection with x-rays Subject RIV: JJ - Other Materials Impact factor: 1.220, year: 2016 http://iopscience.iop.org/article/10.1088/1748-0221/11/01/C01005/pdf

  11. Development of a fast multi-line x-ray CT detector for NDT

    International Nuclear Information System (INIS)

    Hofmann, T.; Nachtrab, F.; Schlechter, T.; Mühlbauer, J.; Schröpfer, S.; Firsching, M.; Uhlmann, N.; Neubauer, H.; Ernst, J.; Schweiger, T.; Oberst, M.; Meyer, A.

    2015-01-01

    Typical X-ray detectors for non-destructive testing (NDT) are line detectors or area detectors, like e.g. flat panel detectors. Multi-line detectors are currently only available in medical Computed Tomography (CT) scanners. Compared to flat panel detectors, line and multi-line detectors can achieve much higher frame rates. This allows time-resolved 3D CT scans of an object under investigation. Also, an improved image quality can be achieved due to reduced scattered radiation from object and detector themselves. Another benefit of line and multi-line detectors is that very wide detectors can be assembled easily, while flat panel detectors are usually limited to an imaging field with a size of approx. 40 × 40 cm 2 at maximum. The big disadvantage of line detectors is the limited number of object slices that can be scanned simultaneously. This leads to long scan times for large objects. Volume scans with a multi-line detector are much faster, but with almost similar image quality. Due to the promising properties of multi-line detectors their application outside of medical CT would also be very interesting for NDT. However, medical CT multi-line detectors are optimized for the scanning of human bodies. Many non-medical applications require higher spatial resolutions and/or higher X-ray energies. For those non-medical applications we are developing a fast multi-line X-ray detector.In the scope of this work, we present the current state of the development of the novel detector, which includes several outstanding properties like an adjustable curved design for variable focus-detector-distances, conserving nearly uniform perpendicular irradiation over the entire detector width. Basis of the detector is a specifically designed, radiation hard CMOS imaging sensor with a pixel pitch of 200 μ m. Each pixel has an automatic in-pixel gain adjustment, which allows for both: a very high sensitivity and a wide dynamic range. The final detector is planned to have 256 lines of

  12. Combined SPECT/CT and PET/CT for breast imaging

    Energy Technology Data Exchange (ETDEWEB)

    Russo, Paolo [Università di Napoli Federico II, Dipartimento di Fisica, Via Cintia, Naples I-80126 (Italy); INFN Sezione di Napoli, Via Cintia, Naples I-80126 (Italy); Larobina, Michele [Istituto di Biostrutture e Bioimmagini, Consiglio Nazionale delle Ricerche, Via Tommaso De Amicis, 95, Naples I-80145 (Italy); Di Lillo, Francesca [Università di Napoli Federico II, Dipartimento di Fisica, Via Cintia, Naples I-80126 (Italy); INFN Sezione di Napoli, Via Cintia, Naples I-80126 (Italy); Del Vecchio, Silvana [Università di Napoli Federico II, Dipartimento di Scienze Biomediche Avanzate, Via Pansini, 5, Naples I-80131 (Italy); Mettivier, Giovanni, E-mail: mettivier@na.infn.it [Università di Napoli Federico II, Dipartimento di Fisica, Via Cintia, Naples I-80126 (Italy); INFN Sezione di Napoli, Via Cintia, Naples I-80126 (Italy)

    2016-02-11

    In the field of nuclear medicine imaging, breast imaging for cancer diagnosis is still mainly based on 2D imaging techniques. Three-dimensional tomographic imaging with whole-body PET or SPECT scanners, when used for imaging the breast, has performance limits in terms of spatial resolution and sensitivity, which can be overcome only with a dedicated instrumentation. However, only few hybrid imaging systems for PET/CT or SPECT/CT dedicated to the breast have been developed in the last decade, providing complementary functional and anatomical information on normal breast tissue and lesions. These systems are still under development and clinical trials on just few patients have been reported; no commercial dedicated breast PET/CT or SPECT/CT is available. This paper reviews combined dedicated breast PET/CT and SPECT/CT scanners described in the recent literature, with focus on their technological aspects.

  13. Ex-situ time-lapse x-ray CT study of 3D micro-structural fatigue damage evolution in uni-directional composites

    DEFF Research Database (Denmark)

    Jespersen, Kristine Munk; Wang, Ying; Zangenberg Hansen, Jens

    2016-01-01

    In this study, the progress of damage under tension-tension fatigue of a uni-directional (UD) glass fibre composite made from a non-crimp fabric is studied using transilluminated white light imaging (TWLI) and X-ray computed tomography (CT). TWLI images are automatically captured throughout...... to initiate already after the first cycle, whereas some grow gradually and others appear suddenly during cycling. The off-axis cracks are observed to saturate after a few thousand cycles. The UD fibre fracture damage in the region observed by X-ray CT is probably already saturated at the first interruption...... point, as no significant change is seen between the two X-ray images. However, the study indicates how TWLI can be used as an initial indicator to locate damage regions at an early stage for the future ex-situ X-ray CT experiments....

  14. Incoherent-scatter computed tomography with monochromatic synchrotron x ray: feasibility of multi-CT imaging system for simultaneous measurement-of fluorescent and incoherent scatter x rays

    Science.gov (United States)

    Yuasa, T.; Akiba, M.; Takeda, T.; Kazama, M.; Hoshino, A.; Watanabe, Y.; Hyodo, K.; Dilmanian, F. A.; Akatsuka, T.; Itai, Y.

    1997-10-01

    We describe a new system of incoherent scatter computed tomography (ISCT) using monochromatic synchrotron X rays, and we discuss its potential to be used in in vivo imaging for medical use. The system operates on the basis of computed tomography (CT) of the first generation. The reconstruction method for ISCT uses the least squares method with singular value decomposition. The research was carried out at the BLNE-5A bending magnet beam line of the Tristan Accumulation Ring in KEK, Japan. An acrylic cylindrical phantom of 20-mm diameter containing a cross-shaped channel was imaged. The channel was filled with a diluted iodine solution with a concentration of 200 /spl mu/gI/ml. Spectra obtained with the system's high purity germanium (HPGe) detector separated the incoherent X-ray line from the other notable peaks, i.e., the iK/sub /spl alpha// and K/sub /spl beta/1/ X-ray fluorescent lines and the coherent scattering peak. CT images were reconstructed from projections generated by integrating the counts In the energy window centering around the incoherent scattering peak and whose width was approximately 2 keV. The reconstruction routine employed an X-ray attenuation correction algorithm. The resulting image showed more homogeneity than one without the attenuation correction.

  15. X-ray and CT signs of connective tissue dysplasia in patients with primarily diagnosed infiltrative pulmonary tuberculosis

    International Nuclear Information System (INIS)

    Sukhanova, L.A.; Sharmazanova, O.P.

    2009-01-01

    The x-ray signs of connective tissue systemic dysplasia (CTSD) in patients with primarily diagnosed pulmonary tuberculosis was investigated. Fifty-four patients (28 med and 26 women aged 18-70) with primarily diagnosed infiltrative pulmonary tuberculosis underwent x-ray study. In patients with infiltration pulmonary tuberculosis CTSD in the lungs manifests by their diminishing, deformity of the lung pattern, high position of the diaphragm cupola, mediastinum shift to the side of the pathology, which is better seen on CT. The degree of CTSD x-ray signs in the lungs depends on the number of phenotypical signs that is the degree of the disease manifestation. CT allows more accurate determining of the signs of connective tissue dysplasia in which tuberculosis develops

  16. Comparison of X-ray, CT and MRI in detection of abnormal sacroiliac joint changes in patients with early stage of ankylosing spondylitis

    International Nuclear Information System (INIS)

    Huang Zhenguo; Zhang Xuezhe; Hong Wen; Wang Guochun; Zhou Huiqiong; Lu Xin; Wang Wu

    2011-01-01

    Objective: To compare X-ray, CT, and MRI in detection of abnormal sacroiliac joint changes in patients with early stage of ankylosing spondylitis (AS). Methods: Fifty-three patients with clinical suspected early stage of AS underwent X-ray and MRI scan. MR scan sequences for the sacroiliac joints consisted of T 1 -weighted, T 2 -weighted, short time inversion recovery (STIR) and three dimensional balance turbo field echo with water selective excitation (3D-BTFE-WATS) in all patients. In 24 of the patients, fat-saturated contrast-enhanced T 1 -weighted was used. Twenty-five of 53 patients underwent CT scan. The Chi-square test was used to analyse the uniformity of bone erosions detected by X-ray, CT, and MRI. Results: Of the 106 sacroiliac joints in 53 patients, 16 sacroiliac joints with bone erosions were detected by X-ray and 63 sacroiliac joints by MRI. Of the 50 sacroiliac joints in 25 patients, 26 sacroiliac joints with bone erosions were found by CT. With regard to the detection of bone erosions, there was no difference between Cf and MRI (χ 2 =0.16, P>0.05) and there was significant difference between CT and X-ray or MRI and X-ray (χ 2 =14.44 and 17.36, P<0.05). 3D-BTFE-WATS was better than other sequences in detection of bone erosions. Acute inflammatory changes were determined by MRI, which included subchondral bone marrow edema in 32 patients, synovitis in 35 patients, fat depositions in 16 patients, enthesitis in 15 patients, capsulitis in 9 patients, and cartilaginous disruption in 31 patients. Conclusions: MRI can detect acute inflammatory changes that can not display by X-ray and CT. Compared with radiography and CT, MRI is more useful in detection of abnormal sacroiliac joint changes in patients with early stage of AS. (authors)

  17. Soil pore characteristics assessed from X-ray micro-CT derived images and correlations to soil friability

    DEFF Research Database (Denmark)

    Munkholm, Lars Juhl; Heck, R; Deen, B

    2012-01-01

    X-ray computed tomography (CT) scanning technology has, in recent decades, been shown to be a very powerful technique to visualize and quantify soil structure. The objective of this project was to quantify soilporecharacteristics, on undisturbed field moist soil, using a high resolution X-rayCT...... cores varied in porosity and porecharacteristics. A drop shatter test was used as a reference procedure to quantify soilfriability. The top 40 mm of the 80 mm high soil samples were scanned using a X-raymicro-CT scanner. The selected region of interest (36 × 36 × 36 mm) was reconstructed with a voxel...... of 49 m3 100 m− 3. The air-filled porosity, at sampling/testing, ranged between 5 and 32 m3 100 m− 3, with an average of 15 m3 100 m− 3. The porosity determined from CT imagery ranged between 1 and 31 m3 100 m− 3, with an average of 4.5 m3 100 m− 3. The number of branches, junctions and end points...

  18. Use of a priori information in incomplete data x-ray CT imaging

    International Nuclear Information System (INIS)

    Eberhard, J.W.; Hedengren, K.H.

    1988-01-01

    A new technique for utilizing a priori information is presented which uses CAD electronic part models to make use of effectively all the information which is available in the blueprint of a selected industrial part. Significant improvements in x-ray image quality are demonstrated using the technique in the image enhancement of the model of an exhaust nozzle actuation ring for the F110 aircraft. Three approaches were evaluated: a projection data approach, an iterative reconstruction approach, and an image processing and analysis approach. Results for these approaches are included. X-ray CT images of the simulated part image reconstructed with several choices of available angular range are shown

  19. Development of a Method to Assess the Precision Of the z-axis X-ray Beam Collimation in a CT Scanner

    Science.gov (United States)

    Kim, Yon-Min

    2018-05-01

    Generally X-ray equipment specifies the beam collimator for the accuracy measurement as a quality control item, but the computed tomography (CT) scanner with high dose has no collimator accuracy measurement item. If the radiation dose is to be reduced, an important step is to check if the beam precisely collimates at the body part for CT scan. However, few ways are available to assess how precisely the X-ray beam is collimated. In this regard, this paper provides a way to assess the precision of z-axis X-ray beam collimation in a CT scanner. After the image plate cassette had been exposed to the X-ray beam, the exposed width was automatically detected by using a computer program developed by the research team to calculate the difference between the exposed width and the imaged width (at isocenter). The result for the precision of z-axis X-ray beam collimation showed that the exposed width was 3.8 mm and the overexposure was high at 304% when a narrow beam of a 1.25 mm imaged width was used. In this study, the precision of the beam collimation of the CT scanner, which is frequently used for medical services, was measured in a convenient way by using the image plate (IP) cassette.

  20. Tomographic image reconstruction using x-ray phase information

    Science.gov (United States)

    Momose, Atsushi; Takeda, Tohoru; Itai, Yuji; Hirano, Keiichi

    1996-04-01

    We have been developing phase-contrast x-ray computed tomography (CT) to make possible the observation of biological soft tissues without contrast enhancement. Phase-contrast x-ray CT requires for its input data the x-ray phase-shift distributions or phase-mapping images caused by an object. These were measured with newly developed fringe-scanning x-ray interferometry. Phase-mapping images at different projection directions were obtained by rotating the object in an x-ray interferometer, and were processed with a standard CT algorithm. A phase-contrast x-ray CT image of a nonstained cancerous tissue was obtained using 17.7 keV synchrotron x rays with 12 micrometer voxel size, although the size of the observation area was at most 5 mm. The cancerous lesions were readily distinguishable from normal tissues. Moreover, fine structures corresponding to cancerous degeneration and fibrous tissues were clearly depicted. It is estimated that the present system is sensitive down to a density deviation of 4 mg/cm3.

  1. Odontogenic maxillary sinusitis diagnosed using conebeam x-ray CT

    International Nuclear Information System (INIS)

    Sato, Kiminori

    2007-01-01

    We evaluated the usefulness of conebeam x-ray CT in the diagnosis of odontogenic maxillary sinusitis in 21 patients. Among teeth causing odontogenic maxillary sinusitis, 95% had apical lesions after root canal treatment. Most root canals were filled with filling materials incompletely. Apical lesions in inappropriately treated teeth thus caused odontogenic maxillary sinusitis. Conebeam CT involves 3-dimensional isotropic voxel image date in up to 512 frames for transaxial, coronal, and sagittal planes, so resolution in imaging on the body axis was especially high. Multiplanar reconstruction and volume rendering images at any optional plane could be obtained without interpolation. The relationship between causative teeth and the maxillary sinus could be observed and measured, and odontogenic maxillary sinusitis accurately diagnosed. In addition to the accurate diagnosis of apical lesions, maxilla, and maxillary sinus, periodontal ligament space, lamina dura, pulp cavity, root canal, canal-treated root, apical periodontitis, alveolar ostitis, marginal periodontitis of causative teeth could be observed. Metal artifacts were minimized, making conebeam CT useful in the diagnosis of periodontal tissue and causative teeth, including root-canal-treated and crown-restored teeth. (author)

  2. Low-dose x-ray phase-contrast and absorption CT using equally sloped tomography

    International Nuclear Information System (INIS)

    Fahimian, Benjamin P; Miao Jianwei; Mao Yu; Cloetens, Peter

    2010-01-01

    Tomographic reconstruction from undersampled and noisy projections is often desirable in transmission CT modalities for purposes of low-dose tomography and fast acquisition imaging. However under such conditions, due to the violation of the Nyquist sampling criteria and the presence of noise, reconstructions with acceptable accuracy may not be possible. Recent experiments in transmission electron tomography and coherent diffraction microscopy have shown that the technique of equally sloped tomography (EST), an exact tomographic method utilizing an oversampling iterative Fourier-based reconstruction, provides more accurate image reconstructions when the number of projections is significantly undersampled relative to filtered back projection and algebraic iterative methods. Here we extend this technique by developing new reconstruction algorithms which allow for the incorporation of advanced mathematical regularization constraints, such as the nonlocal means total variational model, in a manner that is consistent with experimental projections. We then evaluate the resulting image quality of the developed algorithm through simulations and experiments at the European Synchrotron Radiation Facility on image quality phantoms using the x-ray absorption and phase contrast CT modalities. Both our simulation and experimental results have indicated that the method can reduce the number of projections by 60-75% in parallel beam modalities, while achieving comparable or better image quality than the conventional reconstructions. As large-scale and compact synchrotron radiation facilities are currently under rapid development worldwide, the implementation of low-dose x-ray absorption and phase-contrast CT can find broad applications in biology and medicine using these advanced x-ray sources.

  3. WE-FG-207A-02: Why We Need Breast CT? - Clinical Perspective

    International Nuclear Information System (INIS)

    O’Connell, A.

    2016-01-01

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  4. WE-FG-207A-01: Introduction to Dedicated Breast CT - Early Studies

    International Nuclear Information System (INIS)

    Vedantham, S.

    2016-01-01

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  5. WE-FG-207A-02: Why We Need Breast CT? - Clinical Perspective

    Energy Technology Data Exchange (ETDEWEB)

    O’Connell, A. [University of Rochester Medical Center (United States)

    2016-06-15

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  6. WE-FG-207A-04: Performance Characteristics of Photon-Counting Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Kalender, W. [University of Erlangen (Germany)

    2016-06-15

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  7. WE-FG-207A-01: Introduction to Dedicated Breast CT - Early Studies

    Energy Technology Data Exchange (ETDEWEB)

    Vedantham, S. [University of Massachusetts Medical School (United States)

    2016-06-15

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  8. Study on optical properties of ceramics scintillator for X-ray CT equipment. (Pt. 2)

    International Nuclear Information System (INIS)

    Nakamura, Ryouhei; Tsukuda, Yasuo; Nitanda, Humio

    1993-01-01

    The scintillator elements in X-ray CT equipment of the third generation type require high uniformity of distribution of light output as well as quality response characteristic. Parameters responsible for the distribution of light output were studied using the Monte Carlo method. Although the scattering coefficient was neglected, close agreement was obtained between calculated and experimental results. Calculated results indicated that relative output drop at both ends decreased with increasing optical absorption coefficient and was constant for X-ray absorption factor. (author)

  9. Sub-mSV breast XACT scanner: concept and design

    Science.gov (United States)

    Tang, Shanshan; Ren, Liqiang; Samant, Pratik; Chen, Jian; Liu, Hong; Xiang, Liangzhong

    2016-04-01

    Excessive exposure to radiation increases the risk of cancer. We present the concept and design of a new imaging paradigm, X-ray induced acoustic computed tomography (XACT). Applying this innovative technology to breast imaging, one single X-ray exposure can generate a 3D acoustic image, which dramatically reduces the radiation dose to patients when compared to beast CT. A theoretical model is developed to analyze the sensitivity of XACT. A noise equivalent pressure model is used for calculating the minimal radiation dose in XACT imaging. Furthermore, K-Wave simulation is employed to study the acoustic wave propagation in breast tissue. Theoretical analysis shows that the X-ray induced acoustic signal has a 100% relative sensitivity to the X-ray absorption (given that the percentage change in the X-ray absorption coefficient yields the same percentage change in the acoustic signal amplitude), but not to X-ray scattering. The final detection sensitivity is primarily limited by the thermal noise. The radiation dose can be reduced by a factor of 100 compared with the newly FDA approved breast CT. Reconstruction result shows that breast calcification with diameter of 80 μm can be observed in XACT image by using ultrasound transducers with 5.5 MHz center frequency. Therefore, with the proposed innovative technology, one can potentially reduce radiation dose to patient in breast imaging as compared with current x-ray modalities.

  10. TH-AB-209-01: Making Benchtop X-Ray Fluorescence Computed Tomography (XFCT) Practical for in Vivo Imaging by Integration of a Dedicated High-Performance X-Ray Source in Conjunction with Micro-CT Functionality

    International Nuclear Information System (INIS)

    Manohar, N; Cho, S; Reynoso, F

    2016-01-01

    Purpose: To make benchtop x-ray fluorescence computed tomography (XFCT) practical for routine preclinical imaging tasks with gold nanoparticles (GNPs) by deploying, integrating, and characterizing a dedicated high-performance x-ray source and addition of simultaneous micro-CT functionality. Methods: Considerable research effort is currently under way to develop a polychromatic benchtop cone-beam XFCT system capable of imaging GNPs by stimulation and detection of gold K-shell x-ray fluorescence (XRF) photons. Recently, an ad hoc high-power x-ray source was incorporated and used to image the biodistribution of GNPs within a mouse, postmortem. In the current work, a dedicated x-ray source system featuring a liquid-cooled tungsten-target x-ray tube (max 160 kVp, ∼3 kW power) was deployed. The source was operated at 125 kVp, 24 mA. The tube’s compact dimensions allowed greater flexibility for optimizing both the irradiation and detection geometries. Incident x-rays were shaped by a conical collimator and filtered by 2 mm of tin. A compact “OEM” cadmium-telluride x-ray detector was implemented for detecting XRF/scatter spectra. Additionally, a flat panel detector was installed to allow simultaneous transmission CT imaging. The performance of the system was characterized by determining the detection limit (10-second acquisition time) for inserts filled with water/GNPs at various concentrations (0 and 0.010–1.0 wt%) and embedded in a small-animal-sized phantom. The phantom was loaded with 0.5, 0.3, and 0.1 wt% inserts and imaged using XFCT and simultaneous micro-CT. Results: An unprecedented detection limit of 0.030 wt% was experimentally demonstrated, with a 33% reduction in acquisition time. The reconstructed XFCT image accurately localized the imaging inserts. Micro-CT imaging did not provide enough contrast to distinguish imaging inserts from the phantom under the current conditions. Conclusion: The system is immediately capable of in vivo preclinical XFCT

  11. TH-AB-209-01: Making Benchtop X-Ray Fluorescence Computed Tomography (XFCT) Practical for in Vivo Imaging by Integration of a Dedicated High-Performance X-Ray Source in Conjunction with Micro-CT Functionality

    Energy Technology Data Exchange (ETDEWEB)

    Manohar, N; Cho, S [UT MD Anderson Cancer Center, Houston, TX (United States); Reynoso, F [UT MD Anderson Cancer Center, Houston, TX (United States); Washington University School of Medicine, St. Louis, MO (United States)

    2016-06-15

    Purpose: To make benchtop x-ray fluorescence computed tomography (XFCT) practical for routine preclinical imaging tasks with gold nanoparticles (GNPs) by deploying, integrating, and characterizing a dedicated high-performance x-ray source and addition of simultaneous micro-CT functionality. Methods: Considerable research effort is currently under way to develop a polychromatic benchtop cone-beam XFCT system capable of imaging GNPs by stimulation and detection of gold K-shell x-ray fluorescence (XRF) photons. Recently, an ad hoc high-power x-ray source was incorporated and used to image the biodistribution of GNPs within a mouse, postmortem. In the current work, a dedicated x-ray source system featuring a liquid-cooled tungsten-target x-ray tube (max 160 kVp, ∼3 kW power) was deployed. The source was operated at 125 kVp, 24 mA. The tube’s compact dimensions allowed greater flexibility for optimizing both the irradiation and detection geometries. Incident x-rays were shaped by a conical collimator and filtered by 2 mm of tin. A compact “OEM” cadmium-telluride x-ray detector was implemented for detecting XRF/scatter spectra. Additionally, a flat panel detector was installed to allow simultaneous transmission CT imaging. The performance of the system was characterized by determining the detection limit (10-second acquisition time) for inserts filled with water/GNPs at various concentrations (0 and 0.010–1.0 wt%) and embedded in a small-animal-sized phantom. The phantom was loaded with 0.5, 0.3, and 0.1 wt% inserts and imaged using XFCT and simultaneous micro-CT. Results: An unprecedented detection limit of 0.030 wt% was experimentally demonstrated, with a 33% reduction in acquisition time. The reconstructed XFCT image accurately localized the imaging inserts. Micro-CT imaging did not provide enough contrast to distinguish imaging inserts from the phantom under the current conditions. Conclusion: The system is immediately capable of in vivo preclinical XFCT

  12. The color of X-rays Spectral X-ray computed tomography using energy sensitive pixel detectors

    CERN Document Server

    Schioppa, Enrico Junior

    Energy sensitive X-ray imaging detectors are produced by connecting a semiconductor sensor to a spectroscopic pixel readout chip. In this thesis, the applicability of such detectors to X-ray Computed Tomography (CT) is studied. A prototype Medipix based silicon detector is calibrated using X-ray fluorescence. The charge transport properties of the sensor are characterized using a high energy beam of charged particles at the Super Proton Synchrotron (SPS) at the European Center for Nuclear Research (CERN). Monochromatic X-rays at the European Synchrotron Radiation Facility (ESRF) are used to determined the energy response function. These data are used to implement a physics-based CT projection operator that accounts for the transmission of the source spectrum through the sample and detector effects. Based on this projection operator, an iterative spectral CT reconstruction algorithm is developed by extending an Ordered Subset Expectation Maximization (OSEM) method. Subsequently, a maximum likelihood based algo...

  13. Characterising and linking X-ray CT derived macroporosity parameters to infiltration in soils with contrasting structures

    DEFF Research Database (Denmark)

    Müller, Karin; Katuwal, Sheela; Young, Iain

    2018-01-01

    with X-ray CT. Kunsat was significantly higher in the Andosol than in the Gleysol at all water potentials, and decreased significantly with depth in both soils. The in situ measurements guided the definition of new macroporosity parameters from the X-ray CT reconstructions. For the Andosol, Kunsat......Soils deliver the regulating ecosystem services of water infiltration and distribution, which can be controlled by macropores. Parameterizing macropore hydraulic properties is challenging due to the lack of direct measurement methods. With tension-disc infiltrometry hydraulic properties near...... saturation can be measured. Differentiating between hydrologically active and non-active pores, at a given water potential, indirectly assesses macropore continuity. Water flow through macropores is controlled by macropore size distribution, tortuosity, and connectivity, which can be directly derived by X...

  14. Diagnosis of ovarian tumors by ultrasonography and x-ray CT

    International Nuclear Information System (INIS)

    Watanabe, Hiromu

    1986-01-01

    A drawback in ultrasonography, an indispensable technique in imaging ovarian malignancies, is that it sometimes shows false positive for benign tumors including dermoid cyst and ovarian endometriosis. To overcome this, the concurrent use of X-Ray CT seems warrantable and this led us to carry out the study summarized below. CT was carried out on 268 of 322 ovarian disease cases which had undergone ultrasonography. The results obtained were as follows. Sensitivity of ultrasonography and CT in ovarian malignancies were 82 % and 88 % respectively. Of the cases subjected to both techniques, those who were deemed false positive were 39 in ultrasonography and 12 in CT. Ultrasonographic imagings of dermoid cyst and ovarian endometriosis were classified on the basis of their patterns. The results obtained revealed that those in which it was difficult to discriminate the image from ovarian malignancy were types V and VI in dermoid cyst and type III-C in ovarian endometriosis. In both dermoid cyst and ovarian endometriosis, nearly a 100 % accuracy rate was obtainable with CT for diagnosis of benign tumors. This strongly suggested that concurrent use of CT was required, in diagnosing ovarian diseases with the V or VI and III-C patterns referred to above. (author)

  15. Unwrapping an Ancient Egyptian Mummy Using X-Rays

    Science.gov (United States)

    Hughes, Stephen W.

    2010-01-01

    This article describes a project of unwrapping an ancient Egyptian mummy using x-ray computed tomography (CT). About 600 x-ray CT images were obtained through the mummified body of a female named Tjetmutjengebtiu (or Jeni for short), who was a singer in the great temple of Karnak in Egypt during the 22nd dynasty (c 945-715 BC). The x-ray CT images…

  16. Cross-sectional imaging of large and dense materials by high energy X-ray CT using linear accelerator

    International Nuclear Information System (INIS)

    Kanamori, Takahiro; Kamata, Shouji; Ito, Shinichi.

    1989-01-01

    A prototype high energy X-ray CT (computed tomography) system has been developed which employs a linear accelerator as the X-ray source (max. photon energy: 12 MeV). One problem encountered in development of this CT system was to reduce the scattered photons from adjacent detectors, i.e. crosstalk, due to high energy X-rays. This crosstalk was reduced to 2% by means of detector shields using tungsten spacers. Spatial resolution was not affected by such small crosstalk as confirmed by numerical simulations. A second problem was to reduce the scattered photons from the test object. This was done using collimators. A third concern was to realize a wide dynamic range data processing which would allow applications to large and dense objects. This problem was solved by using a sample and hold data acquisition method to reduce the dark current of the photo detectors. The dynamic range of this system was experimentally confirmed over 60 dB. It was demonstrated that slits (width: 2 mm) in an iron object (diameter: 25 cm) could be imaged by this prototype CT system. (author)

  17. Development of real-time x-ray microtomography system

    International Nuclear Information System (INIS)

    Takano, H; Morikawa, M; Konishi, S; Azuma, H; Shimomura, S; Tsusaka, Y; Kagoshima, Y; Nakano, S; Kosaka, N; Yamamoto, K

    2013-01-01

    We have developed a four-dimensional (4D) x-ray microcomputed tomography (CT) system that can obtain time-lapse CT volumes in real time. The system consists of a high-speed sample rotation system and a high-frame-rate x-ray imager, which are installed at a synchrotron radiation x-ray beamline. As a result of system optimization and introduction of a 'zoom resolution' procedure, a real-time 4D CT movie with a frame rate of 30 was obtained with a voxel size of 2.5 μm using 10 keV x-rays

  18. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... over time. top of page What are the benefits vs. risks? Benefits Bone x-rays are the fastest and easiest ... bear denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams ...

  19. The color of X-rays: Spectral X-ray computed tomography using energy sensitive pixel detectors

    NARCIS (Netherlands)

    Schioppa, E.J.

    2014-01-01

    Energy sensitive X-ray imaging detectors are produced by connecting a semiconductor sensor to a spectroscopic pixel readout chip. In this thesis, the applicability of such detectors to X-ray Computed Tomography (CT) is studied. A prototype Medipix based silicon detector is calibrated using X-ray

  20. The oblique view for spondylolysis in the growing period. X-ray projection angle to spondylolytic lumbar vertebra based on CT scanning

    Energy Technology Data Exchange (ETDEWEB)

    Kobayashi, Yoshimitsu; Minato, Izumi; Nagano, Junji; Inoue, Yoshiya; Takahashi, Yuji; Saito, Hidehiko [Seirei Hamamatsu General Hospital, Shizuoka (Japan)

    1996-02-01

    In this study, CT images of 239 vertebrae from 227 patients with spondylolysis in the growing period were obtained to analyze x-ray oblique separation images based on the relationship between x-ray incidence angle and CT. The age of the subjects ranged from 5 to 23 years (average 14.8 years), of which 224 patients were 18 years or lower. The CT images obtained were classified into four according to our classification, i.e., Type Ia, 112 vertebrae; Type Ib, 175; Type II, 66; Type III, 67. All patients underwent plain x-ray from 4 directions including 45deg. The angle of separation line and posterior margin of the vertebral body was obtained and the average angle of all slices was determined as separation angle. As a result, depiction level of separation line was better in 30deg oblique images than 45deg oblique images in Ia, Ib, and III groups. It is thus desirable to take the relationship between incidence angle of x-ray, separation angle, and CT classification into consideration. (S.Y.).

  1. The oblique view for spondylolysis in the growing period. X-ray projection angle to spondylolytic lumbar vertebra based on CT scanning

    International Nuclear Information System (INIS)

    Kobayashi, Yoshimitsu; Minato, Izumi; Nagano, Junji; Inoue, Yoshiya; Takahashi, Yuji; Saito, Hidehiko

    1996-01-01

    In this study, CT images of 239 vertebrae from 227 patients with spondylolysis in the growing period were obtained to analyze x-ray oblique separation images based on the relationship between x-ray incidence angle and CT. The age of the subjects ranged from 5 to 23 years (average 14.8 years), of which 224 patients were 18 years or lower. The CT images obtained were classified into four according to our classification, i.e., Type Ia, 112 vertebrae; Type Ib, 175; Type II, 66; Type III, 67. All patients underwent plain x-ray from 4 directions including 45deg. The angle of separation line and posterior margin of the vertebral body was obtained and the average angle of all slices was determined as separation angle. As a result, depiction level of separation line was better in 30deg oblique images than 45deg oblique images in Ia, Ib, and III groups. It is thus desirable to take the relationship between incidence angle of x-ray, separation angle, and CT classification into consideration. (S.Y.)

  2. The methods for detecting multiple small nodules from 3D chest X-ray CT images

    International Nuclear Information System (INIS)

    Hayase, Yosuke; Mekada, Yoshito; Mori, Kensaku; Toriwaki, Jun-ichiro; Natori, Hiroshi

    2004-01-01

    This paper describes a method for detecting small nodules, whose CT values and diameters are more than -600 Hounsfield unit (H.U.) and 2 mm, from three-dimensional chest X-ray CT images. The proposed method roughly consists of two submodules: initial detection of nodule candidates by discriminating between nodule regions and other regions such as blood vessels or bronchi using a shape feature computed from distance values inside the regions and reduction of false positive (FP) regions by using a minimum directional difference filter called minimum directional difference filter (Min-DD) changing its radius suit to the size of the initial candidates. The performance of the proposed method was evaluated by using seven cases of chest X-ray CT images including six abnormal cases where multiple lung cancers are observed. The experimental results for nodules (361 regions in total) showed that sensitivity and FP regions are 71% and 7.4 regions in average per case. (author)

  3. Advanced high speed X-ray CT scanner for measurement and visualization of multi-phase flow

    International Nuclear Information System (INIS)

    Hori, Keiichi; Fujimoto, Tetsuro; Kawanishi, Kohei; Nishikawa, Hideo

    1998-01-01

    The development of an ultra-fast X-ray computed tomography (CT) scanner has been performed. The object of interest is in a transient or unsettled state, which makes the conventional CT scanner inappropriate. A concept of electrical switching of electron beam of X-ray generation unit is adopted to reduce the scanning time instead of a mechanical motion adopted by a conventional CT scanner. The mechanical motion is a major obstacle to improve the scanning speed. A prototype system with a scanning time of 3.6 milliseconds was developed at first. And, the feasibility was confirmed to measure the dynamic events of two-phase flow. However, faster scanning speed is generally required for the practical use in the thermalhydraulics research field. Therefore, the development of advanced type has been performed. This advanced type can operate under the scanning time of 0.5 milliseconds and is applicable for the measurement of the multi-phase flow with velocity up to 4-5 m/s. (author)

  4. Evaluation of physical property of light-weight soil with air foam using X-ray CT method

    International Nuclear Information System (INIS)

    Otani, Jun; Mukunoki, Toshifumi; Kikuchi, Yoshiaki

    2000-01-01

    The objective of this paper is to investigate the physical property of light-weight soil made of dredged slurry mixed with air foam and cement using X-ray CT method. In this study, not only the specimen made in laboratory but also the one sampled at the in-situ construction site were used and the property in the soil was visualized and the distributions of the density an air porosity were evaluated quantitatively using the results of CT scanning. Here, the method of image processing analysis was also used for this evaluation study. Based on the results obtained in this study, it is concluded that the X-ray CT method is a powerful tool even for geotechnical engineering and this makes not only the visualization but also the quantitative ion discussion possible for the light-weight soil with air foam. (author)

  5. Personalized estimates of radiation dose from dedicated breast CT in a diagnostic population and comparison with diagnostic mammography

    International Nuclear Information System (INIS)

    Vedantham, Srinivasan; Shi, Linxi; Karellas, Andrew; O'Connell, Avice M; Conover, David L

    2013-01-01

    This study retrospectively analyzed the mean glandular dose (MGD) to 133 breasts from 132 subjects, all women, who participated in a clinical trial evaluating dedicated breast CT in a diagnostic population. The clinical trial was conducted in adherence to a protocol approved by institutional review boards and the study participants provided written informed consent. Individual estimates of MGD to each breast from dedicated breast CT was obtained by combining x-ray beam characteristics with estimates of breast dimensions and fibroglandular fraction from volumetric breast CT images, and using normalized glandular dose coefficients. For each study participant and for the breast corresponding to that imaged with breast CT, an estimate of the MGD from diagnostic mammography (including supplemental views) was obtained from the DICOM image headers for comparison. This estimate uses normalized glandular dose coefficients corresponding to a breast with 50% fibroglandular weight fraction. The median fibroglandular weight fraction for the study cohort determined from volumetric breast CT images was 15%. Hence, the MGD from diagnostic mammography was corrected to be representative of the study cohort. Individualized estimates of MGD from breast CT ranged from 5.7 to 27.8 mGy. Corresponding to the breasts imaged with breast CT, the MGD from diagnostic mammography ranged from 2.6 to 31.6 mGy. The mean (± inter-breast SD) and the median MGD (mGy) from dedicated breast CT exam were 13.9 ± 4.6 and 12.6, respectively. For the corresponding breasts, the mean (± inter-breast SD) and the median MGD (mGy) from diagnostic mammography were 12.4 ± 6.3 and 11.1, respectively. Statistical analysis indicated that at the 0.05 level, the distributions of MGD from dedicated breast CT and diagnostic mammography were significantly different (Wilcoxon signed ranks test, p = 0.007). While the interquartile range and the range (maximum–minimum) of MGD from dedicated breast CT was lower than

  6. Systematic errors in digital volume correlation due to the self-heating effect of a laboratory x-ray CT scanner

    KAUST Repository

    Wang, B; Pan, B; Tao, Ran; Lubineau, Gilles

    2017-01-01

    The use of digital volume correlation (DVC) in combination with a laboratory x-ray computed tomography (CT) for full-field internal 3D deformation measurement of opaque materials has flourished in recent years. During x-ray tomographic imaging

  7. Suppression of high-density artefacts in x-ray CT images using temporal digital subtraction with application to cryotherapy

    Energy Technology Data Exchange (ETDEWEB)

    Baissalov, R.; Sandison, G.A.; Rewcastle, J.C. [Department of Medical Physics, Tom Baker Cancer Center, Calgary, Canada, T2N 4N2 2 Department of Physics and Astronomy, University of Calgary, Calgary T2N 2N4 (Canada); Donnelly, B.J. [Department of Surgery, Tom Baker Cancer Center, Calgary, Canada, T2N 4N2 4 Department of Surgery, Foothills Hospital, Calgary T2N 2T7 (Canada); Saliken, J.C. [Department of Surgery, Tom Baker Cancer Center, Calgary T2N 4N2 (Canada); Department of Diagnostic Imaging, Foothills Hospital, Calgary T2N 2T7 (Canada); McKinnon, J.G. [Department of Surgery, Foothills Hospital, Calgary T2N 2T7 (Canada); Muldrew, K. [Department of Surgery, Faculty of Medicine, University of Calgary, Calgary T2N 2T7 (Canada)

    2000-05-01

    Image guidance in cryotherapy is usually performed using ultrasound. Although not currently in routine clinical use, x-ray CT imaging is an alternative means of guidance that can display the full 3D structure of the iceball, including frozen and unfrozen regions. However, the quality of x-ray CT images is compromised by the presence of high-density streak artefacts. To suppress these artefacts we applied temporal digital subtraction (TDS). This TDS method has the added advantage of improving the grey-scale contrast between frozen and unfrozen tissue in the CT images. Two sets of CT images were taken of a phantom material, cryoprobes and a urethral warmer (UW) before and during the cryoprobe freeze cycle. The high-density artefacts persisted in both image sets. TDS was performed on these two image sets using the corresponding mask image of unfrozen material and the same geometrical configuration of the cryoprobes and the UW. The resultant difference image had a significantly reduced artefact content. Thus TDS can be used to significantly suppress or eliminate high-density CT streak artefacts without reducing the metallic content of the cryoprobes. In vivo study needs to be conducted to establish the utility of this TDS procedure for CT assisted prostate or liver cryotherapy. Applying TDS in x-ray CT guided cryotherapy will facilitate estimation of the number and location of all frozen and unfrozen regions, potentially making cryotherapy safer and less operator dependent. (author)

  8. Suppression of high-density artefacts in x-ray CT images using temporal digital subtraction with application to cryotherapy

    International Nuclear Information System (INIS)

    Baissalov, R.; Sandison, G.A.; Rewcastle, J.C.; Donnelly, B.J.; Saliken, J.C.; McKinnon, J.G.; Muldrew, K.

    2000-01-01

    Image guidance in cryotherapy is usually performed using ultrasound. Although not currently in routine clinical use, x-ray CT imaging is an alternative means of guidance that can display the full 3D structure of the iceball, including frozen and unfrozen regions. However, the quality of x-ray CT images is compromised by the presence of high-density streak artefacts. To suppress these artefacts we applied temporal digital subtraction (TDS). This TDS method has the added advantage of improving the grey-scale contrast between frozen and unfrozen tissue in the CT images. Two sets of CT images were taken of a phantom material, cryoprobes and a urethral warmer (UW) before and during the cryoprobe freeze cycle. The high-density artefacts persisted in both image sets. TDS was performed on these two image sets using the corresponding mask image of unfrozen material and the same geometrical configuration of the cryoprobes and the UW. The resultant difference image had a significantly reduced artefact content. Thus TDS can be used to significantly suppress or eliminate high-density CT streak artefacts without reducing the metallic content of the cryoprobes. In vivo study needs to be conducted to establish the utility of this TDS procedure for CT assisted prostate or liver cryotherapy. Applying TDS in x-ray CT guided cryotherapy will facilitate estimation of the number and location of all frozen and unfrozen regions, potentially making cryotherapy safer and less operator dependent. (author)

  9. Engineering iodine-doped carbon dots as dual-modal probes for fluorescence and X-ray CT imaging.

    Science.gov (United States)

    Zhang, Miaomiao; Ju, Huixiang; Zhang, Li; Sun, Mingzhong; Zhou, Zhongwei; Dai, Zhenyu; Zhang, Lirong; Gong, Aihua; Wu, Chaoyao; Du, Fengyi

    2015-01-01

    X-ray computed tomography (CT) is the most commonly used imaging technique for noninvasive diagnosis of disease. In order to improve tissue specificity and prevent adverse effects, we report the design and synthesis of iodine-doped carbon dots (I-doped CDs) as efficient CT contrast agents and fluorescence probe by a facile bottom-up hydrothermal carbonization process. The as-prepared I-doped CDs are monodispersed spherical nanoparticles (a diameter of ~2.7 nm) with favorable dispersibility and colloidal stability in water. The aqueous solution of I-doped CDs showed wavelength-dependent excitation and stable photoluminescence similar to traditional carbon quantum dots. Importantly, I-doped CDs displayed superior X-ray attenuation properties in vitro and excellent biocompatibility. After intravenous injection, I-doped CDs were distributed throughout the body and excreted by renal clearance. These findings validated that I-doped CDs with high X-ray attenuation potency and favorable photoluminescence show great promise for biomedical research and disease diagnosis.

  10. The forgotten view: Chest X-ray - Lateral view

    Directory of Open Access Journals (Sweden)

    Abraham M. Ittyachen

    2017-01-01

    Full Text Available With CT (computed tomography chest gaining more importance as a diagnostic tool, chest X-ray especially the lateral view is taken less commonly nowadays. Besides CT chest is also proven to be superior to chest X-ray in patients with major blunt trauma. We are presenting a 68-year old male who was partially treated from outside for a left sided pneumonia. He came to our hospital because of persisting chest pain. Chest X-ray, frontal view (postero-anterior was almost normal except for a mild opacity in the left lower zone. CT scan of the chest revealed a fluid collection posteriorly enclosed within enhancing pleura. Chest X-ray, left lateral view showed a corresponding posterior pleural based opacity. We are presenting this case to highlight the importance of the lateral view of the chest X-ray. In selected cases there is still a role for the lateral view. With the three dimensional visualization provided by the CT, the lateral view of the chest may be easier to understand. Consequent to the initial diagnosis by CT further follow up can be done with the chest X-ray. In a limited way this mitigates unnecessary expenditure and more importantly prevents the patient from exposure to harmful radiation in the form of repeated CT.

  11. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... have special pediatric considerations. The teddy bear denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams Arthritis X-ray, Interventional Radiology and Nuclear Medicine Radiation Safety How to Read Your Radiology Report ...

  12. Visualization of soil particulate organic matter by means of X-ray CT?

    Science.gov (United States)

    Sleutel, Steven; Van Loo, Denis; Maenhout, Peter; Van Hoorebeke, Luc; Cnudde, Veerle; De Neve, Stefaan

    2014-05-01

    The role of soil structure in organic matter (OM) stabilization has been primarily investigated through physical fractionation studies operative at the scale of aggregates and smaller organo-mineral particles. By narrowing down soil structure to an arrangement of mineral and organic particles, the majority of studies did not explore the spatial organization of the soil pore network, the actual habitat of microorganisms. The pore structure of soil can have a significant impact on soil processes like OM decomposition by excluding OM from micro-organisms in small pores, by regulating the diffusion of substrates and metabolites and by regulating aeration and presence of moisture. Because of its ability to visualize the 3D architecture of soil non-destructively, X-ray Computed Tomography (CT) is becoming a widespread tool for studying soil pore network structure. However, phase determination of pore space, soil OM, soil mineral matter (MM) and water is often limited even with the latest technological and software advances, allowing high resolution and better quality imaging. Contrast agents commonly used in histology enable enhancement of X-ray attenuation of targeted structures or compounds. Here we report on the first systematic investigation of the use of such X-ray contrast agents for soil research. An evaluation procedure as well as a method to apply the agents to soil samples was developed and applied on reference soil samples. The effectiveness and selectivity of the contrast agents was evaluated for soil organic matter (SOM), MM and water. Several products were found to selectively increase the attenuation of water or SOM. The four agents with the best OM-staining capabilities (Phosphomolybdenic acid (PMA), silver nitrate, lead nitrate and lead acetate) were further tested on an OM-MM mixture. Observed differences in reactivity of the staining agents with MM components were apparent, suggesting that contrasting agents may have to be selected for the specific

  13. X-ray diagnostics - benefits and risks

    International Nuclear Information System (INIS)

    Bartholomaeus, Melanie

    2016-01-01

    The brochure on benefits and risks of X-ray diagnostics discusses the following issues: X radiation - a pioneering discovery and medical sensation, fundamentals of X radiation, frequency of X-ray examinations in Germany in relation to CT imaging, radiation doses resulting from X-ray diagnostics, benefits of X-ray diagnostics - indication and examples, risks - measures for radiation exposure reductions, avoidance of unnecessary examinations.

  14. X-ray CT evaluation of pulmonary involvements of sarcoidosis

    International Nuclear Information System (INIS)

    Nishimura, Koichi; Izumi, Takateru; Kitaichi, Masanori

    1987-01-01

    We evaluated high resolution CT in 60 patients with histologically diagnosed pulmonary sarcoidosis and, also, studied the relationship between CT and findings in open lung biopsy specimens in 2 cases. The CT findings were as follows: (1) thickening of bronchial wall shadows (27 out of 60 cases, 45.0 %), (2) irregular enlargement of pulmonary vascular shadows (39 cases, 65.0 %), (3) small or large nodular shadows (24 cases, 40.0 %), (4) local volume loss (14 cases, 23.3 %), (5) slightly increased density of localized lung field areas (24 cases, 40.0 %), (6) pleural or subpleural involvement (27 cases, 45.0 %), (7) lymph node enlargement (59 cases, 98.3 %). X-ray CT in 7 patients revealed no evidence of lung field involvement in patients with histologicall confirmed epithelioid cell granuloma in transbronchial lung biopsy specimens. Lesions located within vessels or in the vascular wall, perivascular sheath or alveoli surrounding blood vessels might cause pulmonary vascular shadows to appear swollen on CT. In a comparative study, we found irregular dilatation of pulmonary vascular shadows corresponding to granulomas in the connective tissue sheath of blood vessels. Also, thickening of bronchial wall shadows corresponded to granulomas in and around the bronchial wall. From the point of histopathological view epithelioid cell granulomas in the bronchovascular sheath were most marked in sarcoidosis, and they apperaed on CT as an irregular enlargement of pulmonary vascular shadows and thickening of the bronchial wall. On the other hand, we reported that collapse of alveoli and fibrosis surrounding blood vessels could cause irregular enlargement of pulmonary vascular shadows on CT in idiopathic pulmonary fibrosis (IPF). Such shadows were seen on CT in both sarcoidosis and IPF but the mechanism of their appearance differed. (J.P.N.)

  15. Micro-CTvlab: A web based virtual gallery of biological specimens using X-ray microtomography (micro-CT).

    Science.gov (United States)

    Keklikoglou, Kleoniki; Faulwetter, Sarah; Chatzinikolaou, Eva; Michalakis, Nikitas; Filiopoulou, Irene; Minadakis, Nikos; Panteri, Emmanouela; Perantinos, George; Gougousis, Alexandros; Arvanitidis, Christos

    2016-01-01

    During recent years, X-ray microtomography (micro-CT) has seen an increasing use in biological research areas, such as functional morphology, taxonomy, evolutionary biology and developmental research. Micro-CT is a technology which uses X-rays to create sub-micron resolution images of external and internal features of specimens. These images can then be rendered in a three-dimensional space and used for qualitative and quantitative 3D analyses. However, the online exploration and dissemination of micro-CT datasets are rarely made available to the public due to their large size and a lack of dedicated online platforms for the interactive manipulation of 3D data. Here, the development of a virtual micro-CT laboratory (Micro-CT vlab ) is described, which can be used by everyone who is interested in digitisation methods and biological collections and aims at making the micro-CT data exploration of natural history specimens freely available over the internet. The Micro-CT vlab offers to the user virtual image galleries of various taxa which can be displayed and downloaded through a web application. With a few clicks, accurate, detailed and three-dimensional models of species can be studied and virtually dissected without destroying the actual specimen. The data and functions of the Micro-CT vlab can be accessed either on a normal computer or through a dedicated version for mobile devices.

  16. CT/FMT dual-model imaging of breast cancer based on peptide-lipid nanoparticles

    Science.gov (United States)

    Xu, Guoqiang; Lin, Qiaoya; Lian, Lichao; Qian, Yuan; Lu, Lisen; Zhang, Zhihong

    2016-03-01

    Breast cancer is one of the most harmful cancers in human. Its early diagnosis is expected to improve the patients' survival rate. X-ray computed tomography (CT) has been widely used in tumor detection for obtaining three-dimentional information. Fluorescence Molecular Tomography (FMT) imaging combined with near-infrared fluorescent dyes provides a powerful tool for the acquisition of molecular biodistribution information in deep tissues. Thus, the combination of CT and FMT imaging modalities allows us to better differentiate diseased tissues from normal tissues. Here we developed a tumor-targeting nanoparticle for dual-modality imaging based on a biocompatible HDL-mimicking peptide-phospholipid scaffold (HPPS) nanocarrier. By incorporation of CT contrast agents (iodinated oil) and far-infrared fluorescent dyes (DiR-BOA) into the hydrophobic core of HPPS, we obtained the FMT and CT signals simultaneously. Increased accumulation of the nanoparticles in the tumor lesions was achieved through the effect of the tumor-targeting peptide on the surface of nanoparticle. It resulted in excellent contrast between lesions and normal tissues. Together, the abilities to sensitively separate the lesions from adjacent normal tissues with the aid of a FMT/CT dual-model imaging approach make the targeting nanoparticles a useful tool for the diagnostics of breast cancer.

  17. Matching methods evaluation framework for stereoscopic breast x-ray images.

    Science.gov (United States)

    Rousson, Johanna; Naudin, Mathieu; Marchessoux, Cédric

    2016-01-01

    Three-dimensional (3-D) imaging has been intensively studied in the past few decades. Depth information is an important added value of 3-D systems over two-dimensional systems. Special focuses were devoted to the development of stereo matching methods for the generation of disparity maps (i.e., depth information within a 3-D scene). Dedicated frameworks were designed to evaluate and rank the performance of different stereo matching methods but never considering x-ray medical images. Yet, 3-D x-ray acquisition systems and 3-D medical displays have already been introduced into the diagnostic market. To access the depth information within x-ray stereoscopic images, computing accurate disparity maps is essential. We aimed at developing a framework dedicated to x-ray stereoscopic breast images used to evaluate and rank several stereo matching methods. A multiresolution pyramid optimization approach was integrated to the framework to increase the accuracy and the efficiency of the stereo matching techniques. Finally, a metric was designed to score the results of the stereo matching compared with the ground truth. Eight methods were evaluated and four of them [locally scaled sum of absolute differences (LSAD), zero mean sum of absolute differences, zero mean sum of squared differences, and locally scaled mean sum of squared differences] appeared to perform equally good with an average error score of 0.04 (0 is the perfect matching). LSAD was selected for generating the disparity maps.

  18. Improved spatial resolution and lower-dose pediatric CT imaging: a feasibility study to evaluate narrowing the X-ray photon energy spectrum

    Energy Technology Data Exchange (ETDEWEB)

    Benz, Mark G. [Safer Pediatric Imaging and Engineering Horizons International, Lincoln, VT (United States); Benz, Matthew W. [Southboro Medical Group, Southboro, MA (United States); Birnbaum, Steven B. [Dartmouth Hitchcock Clinic Manchester, Department of Radiology, Manchester, NH (United States); Chason, Eric; Sheldon, Brian W. [Brown University, Division of Engineering, Materials Science and Engineering Program, Providence, RI (United States); McGuire, Dale [R and D Manager, C and G Technologies Inc., Jeffersonville, IN (United States)

    2014-08-15

    This feasibility study has shown that improved spatial resolution and reduced radiation dose can be achieved in pediatric CT by narrowing the X-ray photon energy spectrum. This is done by placing a hafnium filter between the X-ray generator and a pediatric abdominal phantom. A CT system manufactured in 1999 that was in the process of being remanufactured was used as the platform for this study. This system had the advantage of easy access to the X-ray generator for modifications to change the X-ray photon energy spectrum; it also had the disadvantage of not employing the latest post-imaging noise reduction iterative reconstruction technology. Because we observed improvements after changing the X-ray photon energy spectrum, we recommend a future study combining this change with an optimized iterative reconstruction noise reduction technique. (orig.)

  19. High-resolution breast tomography at high energy: a feasibility study of phase contrast imaging on a whole breast

    International Nuclear Information System (INIS)

    Sztrókay, A; Schlossbauer, T; Bamberg, F; Reiser, M F; Coan, P; Diemoz, P C; Brun, E; Bravin, A; Mayr, D

    2012-01-01

    Previous studies on phase contrast imaging (PCI) mammography have demonstrated an enhancement of breast morphology and cancerous tissue visualization compared to conventional imaging. We show here the first results of the PCI analyser-based imaging (ABI) in computed tomography (CT) mode on whole and large (>12 cm) tumour-bearing breast tissues. We demonstrate in this work the capability of the technique of working at high x-ray energies and producing high-contrast images of large and complex specimens. One entire breast of an 80-year-old woman with invasive ductal cancer was imaged using ABI-CT with monochromatic 70 keV x-rays and an area detector of 92×92 µm 2 pixel size. Sagittal slices were reconstructed from the acquired data, and compared to corresponding histological sections. Comparison with conventional absorption-based CT was also performed. Five blinded radiologists quantitatively evaluated the visual aspects of the ABI-CT images with respect to sharpness, soft tissue contrast, tissue boundaries and the discrimination of different structures/tissues. ABI-CT excellently depicted the entire 3D architecture of the breast volume by providing high-resolution and high-contrast images of the normal and cancerous breast tissues. These results are an important step in the evolution of PCI-CT towards its clinical implementation. (paper)

  20. High-resolution breast tomography at high energy: a feasibility study of phase contrast imaging on a whole breast

    Science.gov (United States)

    Sztrókay, A.; Diemoz, P. C.; Schlossbauer, T.; Brun, E.; Bamberg, F.; Mayr, D.; Reiser, M. F.; Bravin, A.; Coan, P.

    2012-05-01

    Previous studies on phase contrast imaging (PCI) mammography have demonstrated an enhancement of breast morphology and cancerous tissue visualization compared to conventional imaging. We show here the first results of the PCI analyser-based imaging (ABI) in computed tomography (CT) mode on whole and large (>12 cm) tumour-bearing breast tissues. We demonstrate in this work the capability of the technique of working at high x-ray energies and producing high-contrast images of large and complex specimens. One entire breast of an 80-year-old woman with invasive ductal cancer was imaged using ABI-CT with monochromatic 70 keV x-rays and an area detector of 92×92 µm2 pixel size. Sagittal slices were reconstructed from the acquired data, and compared to corresponding histological sections. Comparison with conventional absorption-based CT was also performed. Five blinded radiologists quantitatively evaluated the visual aspects of the ABI-CT images with respect to sharpness, soft tissue contrast, tissue boundaries and the discrimination of different structures/tissues. ABI-CT excellently depicted the entire 3D architecture of the breast volume by providing high-resolution and high-contrast images of the normal and cancerous breast tissues. These results are an important step in the evolution of PCI-CT towards its clinical implementation.

  1. X-Ray Psoralen Activated Cancer Therapy (X-PACT)

    Science.gov (United States)

    Oldham, Mark; Yoon, Paul; Fathi, Zak; Beyer, Wayne F.; Adamson, Justus; Liu, Leihua; Alcorta, David; Xia, Wenle; Osada, Takuya; Liu, Congxiao; Yang, Xiao Y.; Dodd, Rebecca D.; Herndon, James E.; Meng, Boyu; Kirsch, David G.; Lyerly, H. Kim; Dewhirst, Mark W.; Fecci, Peter; Walder, Harold; Spector, Neil L.

    2016-01-01

    This work investigates X-PACT (X-ray Psoralen Activated Cancer Therapy): a new approach for the treatment of solid cancer. X-PACT utilizes psoralen, a potent anti-cancer therapeutic with current application to proliferative disease and extracorporeal photopheresis (ECP) of cutaneous T Cell Lymphoma. An immunogenic role for light-activated psoralen has been reported, contributing to long-term clinical responses. Psoralen therapies have to-date been limited to superficial or extracorporeal scenarios due to the requirement for psoralen activation by UVA light, which has limited penetration in tissue. X-PACT solves this challenge by activating psoralen with UV light emitted from novel non-tethered phosphors (co-incubated with psoralen) that absorb x-rays and re-radiate (phosphoresce) at UV wavelengths. The efficacy of X-PACT was evaluated in both in-vitro and in-vivo settings. In-vitro studies utilized breast (4T1), glioma (CT2A) and sarcoma (KP-B) cell lines. Cells were exposed to X-PACT treatments where the concentrations of drug (psoralen and phosphor) and radiation parameters (energy, dose, and dose rate) were varied. Efficacy was evaluated primarily using flow cell cytometry in combination with complimentary assays, and the in-vivo mouse study. In an in-vitro study, we show that X-PACT induces significant tumor cell apoptosis and cytotoxicity, unlike psoralen or phosphor alone (pphosphor, psoralen, or radiation increase. Finally, in an in-vivo pilot study of BALBc mice with syngeneic 4T1 tumors, we show that the rate of tumor growth is slower with X-PACT than with saline or AMT + X-ray (p<0.0001). Overall these studies demonstrate a potential therapeutic effect for X-PACT, and provide a foundation and rationale for future studies. In summary, X-PACT represents a novel treatment approach in which well-tolerated low doses of x-ray radiation are delivered to a specific tumor site to generate UVA light which in-turn unleashes both short- and potentially long

  2. Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose response

    Energy Technology Data Exchange (ETDEWEB)

    Chain, J N M; McAuley, K B [Department of Chemical Engineering, Queen' s University, Kingston, K7L 3N6 (Canada); Jirasek, A [Department of Physics and Astronomy, University of Victoria, Victoria, V8W 3P6 (Canada); Schreiner, L J, E-mail: kim.mcauley@chee.queensu.ca [Cancer Centre of Southeastern Ontario, Kingston, K7L 5P9 (Canada)

    2011-04-07

    This study reports new N-isopropylacrylamide (NIPAM) polymer gel recipes with increased dose sensitivity and improved dose resolution for x-ray CT readout. NIPAM can be used to increase the solubility of N, N'-methylenebisacrylamide (Bis) in aqueous solutions from approximately 3% to 5.5% by weight, enabling the manufacture of dosimeters containing up to 19.5%T, which is the total concentration of NIPAM and Bis by weight. Gelatin is shown to have a mild influence on dose sensitivity when gels are imaged using x-ray CT, and a stronger influence when gels are imaged optically. Phantoms that contain only 3% gelatin and 5 mM tetrakis hydroxymethyl phosphonium chloride are sufficiently stiff for dosimetry applications. The best cosolvent-free gel formulation has a dose sensitivity in the linear range ({approx}0.88 H Gy{sup -1}) that is a small improvement compared to the best NIPAM-based gels that incorporate isopropanol as a cosolvent ({approx}0.80 H Gy{sup -1}). This new gel formulation results in enhanced dose resolution ({approx}0.052 Gy) for x-ray CT readout, making clinical applications of this imaging modality more feasible.

  3. Cosolvent-free polymer gel dosimeters with improved dose sensitivity and resolution for x-ray CT dose response

    International Nuclear Information System (INIS)

    Chain, J N M; McAuley, K B; Jirasek, A; Schreiner, L J

    2011-01-01

    This study reports new N-isopropylacrylamide (NIPAM) polymer gel recipes with increased dose sensitivity and improved dose resolution for x-ray CT readout. NIPAM can be used to increase the solubility of N, N'-methylenebisacrylamide (Bis) in aqueous solutions from approximately 3% to 5.5% by weight, enabling the manufacture of dosimeters containing up to 19.5%T, which is the total concentration of NIPAM and Bis by weight. Gelatin is shown to have a mild influence on dose sensitivity when gels are imaged using x-ray CT, and a stronger influence when gels are imaged optically. Phantoms that contain only 3% gelatin and 5 mM tetrakis hydroxymethyl phosphonium chloride are sufficiently stiff for dosimetry applications. The best cosolvent-free gel formulation has a dose sensitivity in the linear range (∼0.88 H Gy -1 ) that is a small improvement compared to the best NIPAM-based gels that incorporate isopropanol as a cosolvent (∼0.80 H Gy -1 ). This new gel formulation results in enhanced dose resolution (∼0.052 Gy) for x-ray CT readout, making clinical applications of this imaging modality more feasible.

  4. Registration of pencil beam proton radiography data with X-ray CT.

    Science.gov (United States)

    Deffet, Sylvain; Macq, Benoît; Righetto, Roberto; Vander Stappen, François; Farace, Paolo

    2017-10-01

    Proton radiography seems to be a promising tool for assessing the quality of the stopping power computation in proton therapy. However, range error maps obtained on the basis of proton radiographs are very sensitive to small misalignment between the planning CT and the proton radiography acquisitions. In order to be able to mitigate misalignment in postprocessing, the authors implemented a fast method for registration between pencil proton radiography data obtained with a multilayer ionization chamber (MLIC) and an X-ray CT acquired on a head phantom. The registration was performed by optimizing a cost function which performs a comparison between the acquired data and simulated integral depth-dose curves. Two methodologies were considered, one based on dual orthogonal projections and the other one on a single projection. For each methodology, the robustness of the registration algorithm with respect to three confounding factors (measurement noise, CT calibration errors, and spot spacing) was investigated by testing the accuracy of the method through simulations based on a CT scan of a head phantom. The present registration method showed robust convergence towards the optimal solution. For the level of measurement noise and the uncertainty in the stopping power computation expected in proton radiography using a MLIC, the accuracy appeared to be better than 0.3° for angles and 0.3 mm for translations by use of the appropriate cost function. The spot spacing analysis showed that a spacing larger than the 5 mm used by other authors for the investigation of a MLIC for proton radiography led to results with absolute accuracy better than 0.3° for angles and 1 mm for translations when orthogonal proton radiographs were fed into the algorithm. In the case of a single projection, 6 mm was the largest spot spacing presenting an acceptable registration accuracy. For registration of proton radiography data with X-ray CT, the use of a direct ray-tracing algorithm to compute

  5. Acquiring molecular interference functions of X-ray coherent scattering for breast tissues by combination of simulation and experimental methods

    International Nuclear Information System (INIS)

    Chaparian, A.; Oghabian, M. A.; Changizi, V.

    2009-01-01

    Recently, it has been indicated that X-ray coherent scatter from biological tissues can be used to access signature of tissue. Some scientists are interested in studying this effect to get early detection of breast cancer. Since experimental methods for optimization are time consuming and expensive, some scientists suggest using simulation. Monte Carlo codes are the best option for radiation simulation: however, one permanent defect with Monte Carlo codes has been the lack of a sufficient physical model for coherent (Rayleigh) scattering, including molecular interference effects. Materials and Methods: It was decided to obtain molecular interference functions of coherent X-ray scattering for normal breast tissues by combination of modeling and experimental methods. A Monte Carlo simulation program was written to simulate the angular distribution of scattered photons for the normal breast tissue samples. Moreover, experimental diffraction patterns of these tissues were measured by means of energy dispersive X-ray diffraction method. The simulation and experimental data were used to obtain a tabulation of molecular interference functions for breast tissues. Results: With this study a tabulation of molecular interference functions for normal breast tissues Was prepared to facilitate the simulation diffraction patterns of the tissues without any experimental. Conclusion: The method may lead to design new systems for early detection of breast cancer.

  6. Multi-Mounted X-Ray Computed Tomography.

    Science.gov (United States)

    Fu, Jian; Liu, Zhenzhong; Wang, Jingzheng

    2016-01-01

    Most existing X-ray computed tomography (CT) techniques work in single-mounted mode and need to scan the inspected objects one by one. It is time-consuming and not acceptable for the inspection in a large scale. In this paper, we report a multi-mounted CT method and its first engineering implementation. It consists of a multi-mounted scanning geometry and the corresponding algebraic iterative reconstruction algorithm. This approach permits the CT rotation scanning of multiple objects simultaneously without the increase of penetration thickness and the signal crosstalk. Compared with the conventional single-mounted methods, it has the potential to improve the imaging efficiency and suppress the artifacts from the beam hardening and the scatter. This work comprises a numerical study of the method and its experimental verification using a dataset measured with a developed multi-mounted X-ray CT prototype system. We believe that this technique is of particular interest for pushing the engineering applications of X-ray CT.

  7. CT diagnosis in diseases of the breast

    International Nuclear Information System (INIS)

    Han Hongbin; Xie Jingxia

    1998-01-01

    Purpose: To study the CT signs of breast diseases, and discuss the value of CT in the diagnosis of breast cancer and the axillary lymph node (LN) metastases. Materials and methods: Fifty three cases were reported, including breast cancer 30 cases and benign diseases 23 cases. CT were performed in all patients, 44 also had mammography examination. Results: (1) The CT appearance of breast cancer: round or irregular mass, spiculate border, duct retraction, involved Cooper's ligament, deformed adipose space etc. Benign hyperplasia: Irregular mass and symmetrical thickening of breast glands. Cystic hyperplasia: typically multiple, round, liquid-density mass. (2) CT was comparable to mammography in terms of differential diagnosis, however with breast mass located near the axilla or for evaluation of the thoracic wall. CT was superior to mammography. CT was also helpful in detecting LN metastases. Conclusion: CT is valuable in detecting and differentiating breast diseases as well as in the diagnosis of LN metastases

  8. Fourier domain image fusion for differential X-ray phase-contrast breast imaging

    International Nuclear Information System (INIS)

    Coello, Eduardo; Sperl, Jonathan I.; Bequé, Dirk; Benz, Tobias; Scherer, Kai; Herzen, Julia; Sztrókay-Gaul, Anikó; Hellerhoff, Karin; Pfeiffer, Franz; Cozzini, Cristina; Grandl, Susanne

    2017-01-01

    X-Ray Phase-Contrast (XPC) imaging is a novel technology with a great potential for applications in clinical practice, with breast imaging being of special interest. This work introduces an intuitive methodology to combine and visualize relevant diagnostic features, present in the X-ray attenuation, phase shift and scattering information retrieved in XPC imaging, using a Fourier domain fusion algorithm. The method allows to present complementary information from the three acquired signals in one single image, minimizing the noise component and maintaining visual similarity to a conventional X-ray image, but with noticeable enhancement in diagnostic features, details and resolution. Radiologists experienced in mammography applied the image fusion method to XPC measurements of mastectomy samples and evaluated the feature content of each input and the fused image. This assessment validated that the combination of all the relevant diagnostic features, contained in the XPC images, was present in the fused image as well.

  9. Fourier domain image fusion for differential X-ray phase-contrast breast imaging

    Energy Technology Data Exchange (ETDEWEB)

    Coello, Eduardo, E-mail: eduardo.coello@tum.de [GE Global Research, Garching (Germany); Lehrstuhl für Informatikanwendungen in der Medizin & Augmented Reality, Institut für Informatik, Technische Universität München, Garching (Germany); Sperl, Jonathan I.; Bequé, Dirk [GE Global Research, Garching (Germany); Benz, Tobias [Lehrstuhl für Informatikanwendungen in der Medizin & Augmented Reality, Institut für Informatik, Technische Universität München, Garching (Germany); Scherer, Kai; Herzen, Julia [Lehrstuhl für Biomedizinische Physik, Physik-Department & Institut für Medizintechnik, Technische Universität München, Garching (Germany); Sztrókay-Gaul, Anikó; Hellerhoff, Karin [Institute for Clinical Radiology, Ludwig-Maximilians-University Hospital, Munich (Germany); Pfeiffer, Franz [Lehrstuhl für Biomedizinische Physik, Physik-Department & Institut für Medizintechnik, Technische Universität München, Garching (Germany); Cozzini, Cristina [GE Global Research, Garching (Germany); Grandl, Susanne [Institute for Clinical Radiology, Ludwig-Maximilians-University Hospital, Munich (Germany)

    2017-04-15

    X-Ray Phase-Contrast (XPC) imaging is a novel technology with a great potential for applications in clinical practice, with breast imaging being of special interest. This work introduces an intuitive methodology to combine and visualize relevant diagnostic features, present in the X-ray attenuation, phase shift and scattering information retrieved in XPC imaging, using a Fourier domain fusion algorithm. The method allows to present complementary information from the three acquired signals in one single image, minimizing the noise component and maintaining visual similarity to a conventional X-ray image, but with noticeable enhancement in diagnostic features, details and resolution. Radiologists experienced in mammography applied the image fusion method to XPC measurements of mastectomy samples and evaluated the feature content of each input and the fused image. This assessment validated that the combination of all the relevant diagnostic features, contained in the XPC images, was present in the fused image as well.

  10. Systematic Errors in Dimensional X-ray Computed Tomography

    DEFF Research Database (Denmark)

    that it is possible to compensate them. In dimensional X-ray computed tomography (CT), many physical quantities influence the final result. However, it is important to know which factors in CT measurements potentially lead to systematic errors. In this talk, typical error sources in dimensional X-ray CT are discussed...

  11. High energy X-ray CT system using a linear accelerator for automobile parts inspection

    International Nuclear Information System (INIS)

    Kanamori, T.; Sukita, T.

    1995-01-01

    A high energy X-ray CT system (maximum photon energy: 0.95 MeV) has been developed for industrial use. This system employs a linear accelerator as an X-ray source. It is able to image the cross section of automobile parts and can be applied to a solidification analysis study of the cylinder head in an automobile. This paper describes the features of the system and application results which can be related to solidification analysis of the cylinder head when fabricated from an aluminum casting. Some cross-sectional images are also presented as evidence for nondestructive inspection of automobile parts. (orig.)

  12. Empirical average-case relation between undersampling and sparsity in X-ray CT

    DEFF Research Database (Denmark)

    Jørgensen, Jakob Sauer; Sidky, Emil Y.; Hansen, Per Christian

    2015-01-01

    In X-ray computed tomography (CT) it is generally acknowledged that reconstruction methods exploiting image sparsity allow reconstruction from a significantly reduced number of projections. The use of such reconstruction methods is inspired by recent progress in compressed sensing (CS). However......, the CS framework provides neither guarantees of accurate CT reconstruction, nor any relation between sparsity and a sufficient number of measurements for recovery, i.e., perfect reconstruction from noise-free data. We consider reconstruction through 1-norm minimization, as proposed in CS, from data...... obtained using a standard CT fan-beam sampling pattern. In empirical simulation studies we establish quantitatively a relation between the image sparsity and the sufficient number of measurements for recovery within image classes motivated by tomographic applications. We show empirically that the specific...

  13. EVALUATION OF REINFORCING EFFECT ON FACEBOLTS FOR TUNNELING USING X-RAY CT AND CENTRIFUGE MODEL TEST

    Science.gov (United States)

    Takano, Daiki; Otani, Jun; Date, Kensuke; Yokot, Yasuhiro; Nagatani, Hideki

    The purpose of this paper is firstly to simulate the tunnel face failure in laboratory with four cases of model tests by pulling out tunnel model from a sandy ground that are without using auxiliary method nor facebolts and using facebolts with three different lengths of bolts, and secondary, to investigate the behavior of model ground using X-ray computed tomography (CT) scanner to visualize the failure zone in three dimensions. In addition to those results, a series of centrifuge model tests are conducted to confirm the results of X-ray CT test and also to discuss the ground behavior under full scale stress level. Finally, the effect of face bolting method is evaluated based on all the test results.

  14. A comparison of serial MRI with X-ray CT in a patient with Reye-like syndrome

    International Nuclear Information System (INIS)

    Fujii, Yasushi; Kuriyama, Masanori; Yoshimoto, Masahiro; Konishi, Yukuo; Fujisawa, Shinichi; Sudo, Masakatsu; Hayakawa, Katsumi

    1988-01-01

    Serial MRI and X-ray CT scans were performed on a one-year 7-month old girl with Reye-like syndrome. X-ray CT scan revealed marked hemispheric brain edema in the acute stage and moderate ventricular dilatation with diffuse low density area and cortical atrophy in the chronic stage. MRI performed in the chronic stage showed marked progressive changes. Linear high intensity areas were demonstrated on T 2 weighted MRI performed 21 days after the onset. These high intensity areas increased gradually and 60 days after the onset these high intensity areas spread to the grey matter. These high intensity changes seemed to be correspond to cortical necrosis, and those in the white matter to gliosis or degeneration of myelin. (author)

  15. Technical feasibility proof for high-resolution low-dose photon-counting CT of the breast

    Energy Technology Data Exchange (ETDEWEB)

    Kalender, Willi A.; Kolditz, Daniel; Lueck, Ferdinand [University of Erlangen-Nuernberg, Institute of Medical Physics (IMP), Erlangen (Germany); CT Imaging GmbH, Erlangen (Germany); Steiding, Christian [University of Erlangen-Nuernberg, Institute of Medical Physics (IMP), Erlangen (Germany); CT Imaging GmbH, Erlangen (Germany); University Hospital of Erlangen, Institute of Radiology, Erlangen (Germany); Ruth, Veikko; Roessler, Ann-Christin [University of Erlangen-Nuernberg, Institute of Medical Physics (IMP), Erlangen (Germany); Wenkel, Evelyn [University Hospital of Erlangen, Institute of Radiology, Erlangen (Germany)

    2017-03-15

    X-ray computed tomography (CT) has been proposed and evaluated multiple times as a potentially alternative method for breast imaging. All efforts shown so far have been criticized and partly disapproved because of their limited spatial resolution and higher patient dose when compared to mammography. Our concept for a dedicated breast CT (BCT) scanner therefore aimed at novel apparatus and detector design to provide high spatial resolution of about 100 μm and average glandular dose (AGD) levels of 5 mGy or below. Photon-counting technology was considered as a solution to reach these goals. The complete concept was previously evaluated and confirmed by simulations and basic experiments on laboratory setups. We here present measurements of dose, technical image quality parameters and surgical specimen results on such a scanner. For comparison purposes, the specimens were also imaged with digital mammography (DM) and breast tomosynthesis (BT) apparatus. Results show that photon-counting BCT (pcBCT) at 5 mGy AGD offers sufficiently high 3D spatial resolution for reliable detectability of calcifications and soft tissue delineation. (orig.)

  16. Analysis on breast X-ray of chronic schizophrenia and Iiterature review

    International Nuclear Information System (INIS)

    Dang Lianrong; Wang Xuqiang; Xie Junwei; Pei Genxiang

    2008-01-01

    Objective: To research the symptom of breast X-ray of chronic schizophrenia and the influence of its clinical features. Methods: Take 50 patients who have been under treatment in our hospital for over 5 years as sample of psychotic group and same number of healthy people as control group. Analyze their X-ray films, using statistics to deal with their lung area of both before and after 5 years. Evaluation Criteria of X-ray: (1) Double layer rise; (2)Single or double lung base subsegmental atelectasis; (3)Heart drown; (4)Heart turing with the heart apex extending to the left; (5)The double lung bottom have gathering stripes; (6)The upper layer have wider shadow. Results: The lung lobes of the patient group become smaller than they were 5 years ago. Data show great difference between two years (t=6.7099, P<0.01). The lung lobes of psychotic group become bigger than the control group in the first X-ray films (t=2.025, P<0.05), but smaller in the last X-ray films (t=5.512, P<0.01). Among them, 37 cases show double layer rise; 19 cases show single or double lung base subsegmental atelectasis; 24 cases show heart drown; 19 cases show heart turing and heart apex extending to the left; 37 cases show double lung bottom gathering stripes; 17 cases show wider shadow. Conclusion: Morphology and pathology changes of schizophrenia patients are related to the side effect of anti-schizophrenia medicine and endocrine disorders. (authors)

  17. Characterization of a Test for Invasive Breast Cancer Using X-ray Diffraction of Hair - Results of a Clinical Trial

    Directory of Open Access Journals (Sweden)

    Gary L. Corino

    2009-11-01

    Full Text Available Objective: To assess the performance of a test for breast cancer utilizing synchrotron x-ray diffraction analysis of scalp hair from women undergoing diagnostic radiology assessment. Design and Setting: A double-blinded clinical trial of women who attended diagnostic radiology clinics in Australia. Patients: 1796 women referred for diagnostic radiology, with no previous history of cancer. Main Outcome Measures: Sensitivity, specificity and accuracy of the hair test analysis compared to the gold standard of imaging followed by biopsy where indicated. Results: The hair-based assay had an overall accuracy of >77% and a negative predictive value of 99%. For all women, the sensitivity of both mammography and x-ray diffraction alone was 64%, but when used together the sensitivity rose to 86%. The sensitivity of the hair test for women under the age of 70 was 74%. Conclusion: In this large population trial the association between the presence of breast cancer and an altered hair fibre X-ray diffraction pattern previously reported has been confirmed. It appears that mammography and X-ray diffraction of hair detect different populations of breast cancers, and are synergistic when used together.

  18. TU-G-207-03: High Spatial Resolution and High Sensitivity X-Ray Fluorescence Imaging

    International Nuclear Information System (INIS)

    Xing, L.

    2015-01-01

    Last few years has witnessed the development of novel of X-ray imaging modalities, such as spectral CT, phase contrast CT, and X-ray acoustic/fluorescence/luminescence imaging. This symposium will present the recent advances of these emerging X-ray imaging modalities and update the attendees with knowledge in various related topics, including X-ray photon-counting detectors, X-ray physics underlying the emerging applications beyond the traditional X-ray imaging, image reconstruction for the novel modalities, characterization and evaluation of the systems, and their practical implications. In addition, the concept and practical aspects of X-ray activatable targeted nanoparticles for molecular X-ray imaging will be discussed in the context of X-ray fluorescence and luminescence CT. Learning Objectives: Present background knowledge of various emerging X-ray imaging techniques, such as spectral CT, phase contrast CT and X-ray fluorescence/luminescence CT. Discuss the practical need, technical aspects and current status of the emerging X-ray imaging modalities. Describe utility and future impact of the new generation of X-ray imaging applications

  19. A measurement-based X-ray source model characterization for CT dosimetry computations.

    Science.gov (United States)

    Sommerville, Mitchell; Poirier, Yannick; Tambasco, Mauro

    2015-11-08

    The purpose of this study was to show that the nominal peak tube voltage potential (kVp) and measured half-value layer (HVL) can be used to generate energy spectra and fluence profiles for characterizing a computed tomography (CT) X-ray source, and to validate the source model and an in-house kV X-ray dose computation algorithm (kVDoseCalc) for computing machine- and patient-specific CT dose. Spatial variation of the X-ray source spectra of a Philips Brilliance and a GE Optima Big Bore CT scanner were found by measuring the HVL along the direction of the internal bow-tie filter axes. Third-party software, Spektr, and the nominal kVp settings were used to generate the energy spectra. Beam fluence was calculated by dividing the integral product of the spectra and the in-air NIST mass-energy attenuation coefficients by in-air dose measurements along the filter axis. The authors found the optimal number of photons to seed in kVDoseCalc to achieve dose convergence. The Philips Brilliance beams were modeled for 90, 120, and 140 kVp tube settings. The GE Optima beams were modeled for 80, 100, 120, and 140 kVp tube settings. Relative doses measured using a Capintec Farmer-type ionization chamber (0.65 cc) placed in a cylindrical polymethyl methacrylate (PMMA) phantom and irradiated by the Philips Brilliance, were compared to those computed with kVDoseCalc. Relative doses in an anthropomorphic thorax phantom (E2E SBRT Phantom) irradiated by the GE Optima were measured using a (0.015 cc) PTW Freiburg ionization chamber and compared to computations from kVDoseCalc. The number of photons required to reduce the average statistical uncertainty in dose to measurement over all 12 PMMA phantom positions was found to be 1.44%, 1.47%, and 1.41% for 90, 120, and 140 kVp, respectively. The maximum percent difference between calculation and measurement for all energies, measurement positions, and phantoms was less than 3.50%. Thirty-five out of a total of 36 simulation conditions were

  20. Phase-contrast x-ray computed tomography for observing biological specimens and organic materials

    Science.gov (United States)

    Momose, Atsushi; Takeda, Tohoru; Itai, Yuji

    1995-02-01

    A novel three-dimensional x-ray imaging method has been developed by combining a phase-contrast x-ray imaging technique with x-ray computed tomography. This phase-contrast x-ray computed tomography (PCX-CT) provides sectional images of organic specimens that would produce absorption-contrast x-ray CT images with little contrast. Comparing PCX-CT images of rat cerebellum and cancerous rabbit liver specimens with corresponding absorption-contrast CT images shows that PCX-CT is much more sensitive to the internal structure of organic specimens.

  1. Synchrotron radiation X-ray microfluorescence techniques

    Indian Academy of Sciences (India)

    Synchrotron X-ray imaging systems with fluorescence techniques was developed for biomedical researches in Brazilian Synchrotron Laboratory. An X-ray fluorescence microtomography system was implemented to analyse human prostate and breast samples and an X-ray microfluorescence system was implemented to ...

  2. Correlation of conventional simulation x-ray films and CT images for HDR-brachytherapy catheters reconstruction

    International Nuclear Information System (INIS)

    Rajendran, M.; Reddy, K.D.; Reddy, R.M.; Reddy, J.M.; Reddy, B.V.N.; Kiran Kumar; Gopi, S.; Dharaniraj; Janardhanan

    2002-01-01

    In order to plan a brachytherapy implant, it is imperative that implant reconstruction is done accurately. The purpose of this paper is to evaluate whether implant reconstruction done with transverse CT images is comparable to reconstruction done with conventional x-ray films

  3. Prediction of biopore- and matrix-dominated flow from X-ray CT-derived macropore network characteristics

    DEFF Research Database (Denmark)

    Naveed, Muhammad; Møldrup, Per; Schaap, Marcel G

    2016-01-01

    was to investigate the predictive performance of previously developed empirical models for both water and air flow and to explore the potential applicability of X-ray computed tomography (CT)-derived macropore network characteristics. For this purpose, 65 cylindrical soil columns (6 cm diameter and 3.5 cm height......Prediction and modeling of localized flow processes in macropores is of crucial importance for sustaining both soil and water quality. However, currently there are no reliable means to predict preferential flow due to its inherently large spatial variability. The aim of this study......) were extracted from the topsoil (5 cm to 8.5 cm depth) in a 15m15m grid from an agricultural field located in Silstrup, Denmark. All soil columns were scanned with an industrial X-ray CT scanner (129 μm resolution) and later employed for measurement of saturated hydraulic conductivity, air permeability...

  4. Quantitative measurement of lung density with x-ray CT and positron CT, (2)

    International Nuclear Information System (INIS)

    Ito, Kengo; Ito, Masatoshi; Kubota, Kazuo

    1985-01-01

    Lung density was quantitatively measured on six diseased patients with X-ray CT (XCT) and Positron CT(PCT). The findings are as follows: In the silicosis, extravascular lung density was found to be remarkably increased compared to normals (0.29gcm -3 ), but blood volume was in normal range. In the post-irradiated lung cancers, extravascular lung density increased in the irradiated sites compared to the non-irradiated opposite sites, and blood volume varied in each case. In a patient with chronic heart failure, blood volume decreased (0.11mlcm -3 ) with increased extravascular lung density (0.23gcm -3 ). In the chronic obstructive pulmonary disease, both extravascular lung density and blood volume decreased (0.11gcm -3 and 0.10mlcm -3 respectively). Lung density measured with XCT was constantly lower than that with PCT in all cases. But changes in the values of lung density measured, correlated well with each other. In conclusion, the method presented here may clarify the etiology of the diffuse pulmonary diseases, and be used to differentiate and grade the diseases. (author)

  5. A spectral X-ray CT simulation study for quantitative determination of iron

    Science.gov (United States)

    Su, Ting; Kaftandjian, Valérie; Duvauchelle, Philippe; Zhu, Yuemin

    2018-06-01

    Iron is an essential element in the human body and disorders in iron such as iron deficiency or overload can cause serious diseases. This paper aims to explore the ability of spectral X-ray CT to quantitatively separate iron from calcium and potassium and to investigate the influence of different acquisition parameters on material decomposition performance. We simulated spectral X-ray CT imaging of a PMMA phantom filled with iron, calcium, and potassium solutions at various concentrations (15-200 mg/cc). Different acquisition parameters were considered, such as the number of energy bins (6, 10, 15, 20, 30, 60) and exposure factor per projection (0.025, 0.1, 1, 10, 100 mA s). Based on the simulation data, we investigated the performance of two regularized material decomposition approaches: projection domain method and image domain method. It was found that the former method discriminated iron from calcium, potassium and water in all cases and tended to benefit from lower number of energy bins for lower exposure factor acquisition. The latter method succeeded in iron determination only when the number of energy bins equals 60, and in this case, the contrast-to-noise ratios of the decomposed iron images are higher than those obtained using the projection domain method. The results demonstrate that both methods are able to discriminate and quantify iron from calcium, potassium and water under certain conditions. Their performances vary with the acquisition parameters of spectral CT. One can use one method or the other to benefit better performance according to the data available.

  6. Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields.

    Science.gov (United States)

    Kelaranta, Anna; Ekholm, Marja; Toroi, Paula; Kortesniemi, Mika

    2016-01-01

    Dental radiography may involve situations where the patient is known to be pregnant or the pregnancy is noticed after the X-ray procedure. In such cases, the radiation dose to the foetus, though low, needs to be estimated. Uniform and widely used guidance on dental X-ray procedures during pregnancy are presently lacking, the usefulness of lead shields is unclear and practices vary. Upper estimates of radiation doses to the foetus and breasts of the pregnant patient were estimated with an anthropomorphic female phantom in intraoral, panoramic, cephalometric and CBCT dental modalities with and without lead shields. The upper estimates of foetal doses varied from 0.009 to 6.9 μGy, and doses at the breast level varied from 0.602 to 75.4 μGy. With lead shields, the foetal doses varied from 0.005 to 2.1 μGy, and breast doses varied from 0.002 to 10.4 μGy. The foetal dose levels without lead shielding were dental radiographic examination.

  7. Analyses of the internal structure of the oscillating vibro-packed fuels by the micro focus X-rays CT method

    International Nuclear Information System (INIS)

    Mizuta, Yasutoshi

    2003-02-01

    The purpose of this study is to support the development of vibro-packed fuel technology at Japan Nuclear Cycle Development Institute. 3-dimensional (3-D) data was built from the multi-cross sectional images obtained by the micro focus X-rays CT method in the vibro-packed fuel models. The structural analyses were carried out about the obtained 3-D CT images. The packing-rate distribution and the density distribution were measured as well as the number distribution of particles, etc. Consequently, it is obtained that vibrate conditions and a vibrating state have strong correlation, and it is also shown that the 3-D analyses of the internal structure by the micro focus X-rays CT method are effective in performance evaluation of vibro-packed fuels. (author)

  8. Development and prospects of X-ray computerized tomography. I

    International Nuclear Information System (INIS)

    Dobes, V.

    1985-01-01

    The history and developmental trends are described of X-ray computerized tomography (CT) as are its applications in clinical practise. Suitable criteria and economic aspects are proposed for the choice of CT systems. The types are listed and described of X-ray CT systems, including brain, whole-body, mobile and special systems. A table is given showing available specifications (types of detectors, X-ray source, resolution, configuration, scan angles, scan fields, image reconstruction, etc.) for machines by different companies. (M.D.)

  9. Prospective study of the reproducibility of X-rays and CT scans for assessing trochanteric fracture comminution in the elderly: a series of 110 cases.

    Science.gov (United States)

    Isida, Ronald; Bariatinsky, Varenka; Kern, Gregory; Dereudre, Gregoire; Demondion, Xavier; Chantelot, Christophe

    2015-10-01

    Trochanteric fractures are common but difficult to analyse in the elderly on plain X-rays. Fixation failures are related to the severity of the comminution, but comminution and the degree of instability are not easy to determine on standard X-rays. Use of computed tomography (CT) improves assessment of complex fractures, but this finding has not been confirmed versus intraoperative data. The primary objective of this prospective study was to determine the error rate when evaluating comminution on X-rays and CT scans. The secondary objectives were to determine whether CT data on comminution and stability were consistent with intraoperative findings and to define the fracture characteristics. Standard X-ray assessment underestimates the complexity of trochanteric fractures and is not very reproducible. Between January and December 2013, all proximal femur fractures in the trochanter area of patients aged 75 years or older (mean age 85) were analysed prospectively with standard X-rays and computed tomography (CT). One hundred and ten patients (88 women and 22 men) with trochanteric fractures were included in the study. Fracture stability was evaluated using the Müller AO classification; the other fracture characteristics were evaluated independently. A senior surgeon evaluated the anonymised X-rays. A radiologist specialised in musculoskeletal imaging interpreted the CT scan images. All patients underwent total hip arthroplasty (110 patients) and 104 fixations of the greater trochanter. The X-ray and CT findings were compared to the intraoperative findings (gold standard) to evaluate their reproducibility. The reproducibility of the X-ray evaluation was poor for comminution, with a kappa of 0.4, sensitivity of 44 % and a negative predictive value of 29 % but a positive predictive value and specificity of 100 %. The CT evaluation had a kappa of 0.94, sensitivity of 95 % and negative predictive value of 79 %. According to the AO classification, unstable fractures were

  10. An update on carbon nanotube-enabled X-ray sources for biomedical imaging.

    Science.gov (United States)

    Puett, Connor; Inscoe, Christina; Hartman, Allison; Calliste, Jabari; Franceschi, Dora K; Lu, Jianping; Zhou, Otto; Lee, Yueh Z

    2018-01-01

    A new imaging technology has emerged that uses carbon nanotubes (CNT) as the electron emitter (cathode) for the X-ray tube. Since the performance of the CNT cathode is controlled by simple voltage manipulation, CNT-enabled X-ray sources are ideal for the repetitive imaging steps needed to capture three-dimensional information. As such, they have allowed the development of a gated micro-computed tomography (CT) scanner for small animal research as well as stationary tomosynthesis, an experimental technology for large field-of-view human imaging. The small animal CT can acquire images at specific points in the respiratory and cardiac cycles. Longitudinal imaging therefore becomes possible and has been applied to many research questions, ranging from tumor response to the noninvasive assessment of cardiac output. Digital tomosynthesis (DT) is a low-dose and low-cost human imaging tool that captures some depth information. Known as three-dimensional mammography, DT is now used clinically for breast imaging. However, the resolution of currently-approved DT is limited by the need to swing the X-ray source through space to collect a series of projection views. An array of fixed and distributed CNT-enabled sources provides the solution and has been used to construct stationary DT devices for breast, lung, and dental imaging. To date, over 100 patients have been imaged on Institutional Review Board-approved study protocols. Early experience is promising, showing an excellent conspicuity of soft-tissue features, while also highlighting technical and post-acquisition processing limitations that are guiding continued research and development. Additionally, CNT-enabled sources are being tested in miniature X-ray tubes that are capable of generating adequate photon energies and tube currents for clinical imaging. Although there are many potential applications for these small field-of-view devices, initial experience has been with an X-ray source that can be inserted into the

  11. Contrast-enhanced dedicated breast CT detection of invasive breast cancer preceding mammographic diagnosis

    Directory of Open Access Journals (Sweden)

    Nicolas D. Prionas, MD, PhD

    2015-01-01

    Full Text Available Dedicated breast computed tomography (bCT generates high-resolution, three-dimensional images of the pendent uncompressed breast. Intravenous iodinated contrast during bCT provides additional physiologic information. In this case, a 10.0-mm invasive ductal carcinoma was visualized using contrast-enhanced breast CT one year before mammographic detection. Mammography four months before bCT was negative. The bCT contrast enhancement pattern closely matched the dynamic contrast-enhanced MRI obtained after diagnosis. Lesion enhancement at contrast-enhanced breast CT matched previously published enhancement values of breast cancer. Contrast-enhanced dedicated bCT provided high-resolution tomographic images and physiologic contrast enhancement data that facilitated the detection of an early breast cancer.

  12. Effects of x-ray and CT image enhancements on the robustness and accuracy of a rigid 3D/2D image registration

    International Nuclear Information System (INIS)

    Kim, Jinkoo; Yin Fangfang; Zhao Yang; Kim, Jae Ho

    2005-01-01

    A rigid body three-dimensional/two-dimensional (3D/2D) registration method has been implemented using mutual information, gradient ascent, and 3D texturemap-based digitally reconstructed radiographs. Nine combinations of commonly used x-ray and computed tomography (CT) image enhancement methods, including window leveling, histogram equalization, and adaptive histogram equalization, were examined to assess their effects on accuracy and robustness of the registration method. From a set of experiments using an anthropomorphic chest phantom, we were able to draw several conclusions. First, the CT and x-ray preprocessing combination with the widest attraction range was the one that linearly stretched the histograms onto the entire display range on both CT and x-ray images. The average attraction ranges of this combination were 71.3 mm and 61.3 deg in the translation and rotation dimensions, respectively, and the average errors were 0.12 deg and 0.47 mm. Second, the combination of the CT image with tissue and bone information and the x-ray images with adaptive histogram equalization also showed subvoxel accuracy, especially the best in the translation dimensions. However, its attraction ranges were the smallest among the examined combinations (on average 36 mm and 19 deg). Last the bone-only information on the CT image did not show convergency property to the correct registration

  13. Imaging features of breast echinococcus granulosus

    International Nuclear Information System (INIS)

    Zeng Li; Liu Fanming; Gong Yue; Ge Jinmei; Li Xianjun; Shi Minxin; Guo Yongzhong

    2012-01-01

    Objective: To demonstrate the X-ray and CT features of breast hydatid disease. Methods: Of 11 patients with pathologically confirmed breast Echinococcus hydatid disease were collected and the X-ray and CT image data were analyzed. Results: Of 11 patients with hydatid cysts,single cyst was found in 9 patients which one cyst was ruptured due to trauma, multiple cyst in 2 patients. Mammography showed small or large shadow in different size, with low or high density and smooth margin. Calcification was found in 5 and 2 patients with egg shell-like calcification along the wall of cyst, 3 patients with spotted calcification within cyst. One case had cavity-like change (annular solar eclipse sign). Cystic lesion with a complete capsule was demonstrated on CT scan in 1 patient. Conclusion: Molybdenum target mammography can accurately display the imaging characteristics of hydatid cyst and improve the diagnostic ability of breast hydatid cyst in combination with clinical and epidemiological data. (authors)

  14. WE-FG-207A-03: Low-Dose Cone-Beam Breast CT: Physics and Technology Development

    International Nuclear Information System (INIS)

    Boone, J.

    2016-01-01

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  15. WE-FG-207A-03: Low-Dose Cone-Beam Breast CT: Physics and Technology Development

    Energy Technology Data Exchange (ETDEWEB)

    Boone, J. [UC Davis Medical Center (United States)

    2016-06-15

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  16. Visualizing Typical Features of Breast Fibroadenomas Using Phase-Contrast CT: An Ex-Vivo Study

    Science.gov (United States)

    Grandl, Susanne; Willner, Marian; Herzen, Julia; Sztrókay-Gaul, Anikó; Mayr, Doris; Auweter, Sigrid D.; Hipp, Alexander; Birnbacher, Lorenz; Marschner, Mathias; Chabior, Michael; Reiser, Maximilian; Pfeiffer, Franz; Bamberg, Fabian; Hellerhoff, Karin

    2014-01-01

    Background Fibroadenoma is the most common benign solid breast lesion type and a very common cause for histologic assessment. To justify a conservative therapy, a highly specific discrimination between fibroadenomas and other breast lesions is crucial. Phase-contrast imaging offers improved soft-tissue contrast and differentiability of fine structures combined with the potential of 3-dimensional imaging. In this study we assessed the potential of grating-based phase-contrast CT imaging for visualizing diagnostically relevant features of fibroadenomas. Materials and Methods Grating-based phase-contrast CT was performed on six ex-vivo formalin-fixed breast specimens containing a fibroadenoma and three samples containing benign changes that resemble fibroadenomas using Talbot Lau interferometry and a polychromatic X-ray source. Phase-contrast and simultaneously acquired absorption-based 3D-datasets were manually matched with corresponding histological slices. The visibility of diagnostically valuable features was assessed in comparison with histology as the gold-standard. Results In all cases, matching of grating-based phase-contrast CT images and histology was successfully completed. Grating-based phase-contrast CT showed greatly improved differentiation of fine structures and provided accurate depiction of strands of fibrous tissue within the fibroadenomas as well as of the diagnostically valuable dilated, branched ductuli of the fibroadenomas. A clear demarcation of tumor boundaries in all cases was provided by phase- but not absorption-contrast CT. Conclusions Pending successful translation of the technology to a clinical setting and considerable reduction of the required dose, the data presented here suggest that grating-based phase-contrast CT may be used as a supplementary non-invasive diagnostic tool in breast diagnostics. Phase-contrast CT may thus contribute to the reduction of false positive findings and reduce the recall and core biopsy rate in population

  17. A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT

    International Nuclear Information System (INIS)

    Siewerdsen, J.H.; Daly, M.J.; Bakhtiar, B.

    2006-01-01

    X-ray scatter poses a significant limitation to image quality in cone-beam CT (CBCT), resulting in contrast reduction, image artifacts, and lack of CT number accuracy. We report the performance of a simple scatter correction method in which scatter fluence is estimated directly in each projection from pixel values near the edge of the detector behind the collimator leaves. The algorithm operates on the simple assumption that signal in the collimator shadow is attributable to x-ray scatter, and the 2D scatter fluence is estimated by interpolating between pixel values measured along the top and bottom edges of the detector behind the collimator leaves. The resulting scatter fluence estimate is subtracted from each projection to yield an estimate of the primary-only images for CBCT reconstruction. Performance was investigated in phantom experiments on an experimental CBCT benchtop, and the effect on image quality was demonstrated in patient images (head, abdomen, and pelvis sites) obtained on a preclinical system for CBCT-guided radiation therapy. The algorithm provides significant reduction in scatter artifacts without compromise in contrast-to-noise ratio (CNR). For example, in a head phantom, cupping artifact was essentially eliminated, CT number accuracy was restored to within 3%, and CNR (breast-to-water) was improved by up to 50%. Similarly in a body phantom, cupping artifact was reduced by at least a factor of 2 without loss in CNR. Patient images demonstrate significantly increased uniformity, accuracy, and contrast, with an overall improvement in image quality in all sites investigated. Qualitative evaluation illustrates that soft-tissue structures that are otherwise undetectable are clearly delineated in scatter-corrected reconstructions. Since scatter is estimated directly in each projection, the algorithm is robust with respect to system geometry, patient size and heterogeneity, patient motion, etc. Operating without prior information, analytical modeling

  18. Role of X-ray CT in the evaluation of extension of ovarian cancer

    International Nuclear Information System (INIS)

    Saito, Takashi

    1991-01-01

    To evaluate the utility of X-ray computed tomography (CT) in the management of malignant ovarian tumor, preoperative CT of 89 patients with ovarian cancer were analyzed on the basis of their surgically confirmed findings. Relationships between CT findings and clinical factors such as actual tumor spread, stage, and accomplished surgery were studied. The results obtained were as follows. The rates of correct diagnosis in tumor spread were 79.8% for tumor growth on the external surface of the ovary, 77.6% for involvement of both ovaries, 79.8% for ascites, 78.2% for metastasis to the uterus, 68.2% for peritoneal metastasis, 77.0% for metastasis to the intestine, and 79.3% for the metastasis to the omentum. Comparing the stage suggested by CT (CT stage) and the stage revealed by surgery (surgical stage), the stages coincided in 57 patients (64.0%). CT stages were overestimated in 9 caces and underestimated in 23 caces. As for the practicability of surgery, either complete or standard operation was performed in 31 of 33 caces (93.9%) of CT stage I, 21 of 28 (75.0%) of CT stage II, and 11 of 28 (39.3%) of CT stage III. The possibility of accomplishing surgery was very low when CT indicated stage III disease. (author)

  19. Quantitative X-ray microtomography with synchrotron radiation

    Energy Technology Data Exchange (ETDEWEB)

    Donath, T. [GKSS-Forschungszentrum Geesthacht GmbH (Germany). Inst. fuer Materialforschung

    2007-07-01

    Synchrotron-radiation-based computed microtomography (SR{sub {mu}}CT) is an established method for the examination of volume structures. It allows to measure the x-ray attenuation coefficient of a specimen three-dimensionally with a spatial resolution of about one micrometer. In contrast to conventional x-ray sources (x-ray tubes), the unique properties of synchrotron radiation enable quantitative measurements that do not suffer from beam-hardening artifacts. During this work the capabilities for quantitative SR{sub {mu}}CT measurements have been further improved by enhancements that were made to the SR{sub {mu}}CT apparatus and to the reconstruction chain. For high-resolution SR{sub {mu}}CT an x-ray camera consisting of luminescent screen (x-ray phosphor), lens system, and CCD camera was used. A significant suppression of blur that is caused by reflections inside the luminescent screen could be achieved by application of an absorbing optical coating to the screen surface. It is shown that blur and ring artifacts in the tomographic reconstructions are thereby drastically reduced. Furthermore, a robust and objective method for the determination of the center of rotation in projection data (sinograms) is presented that achieves sub-pixel precision. By implementation of this method into the reconstruction chain, complete automation of the reconstruction process has been achieved. Examples of quantitative SR{sub {mu}}CT studies conducted at the Hamburger Synchrotronstrahlungslabor HASYLAB at the Deutsches Elektronen-Synchrotron DESY are presented and used for the demonstration of the achieved enhancements. (orig.)

  20. Quantitative X-ray microtomography with synchrotron radiation

    International Nuclear Information System (INIS)

    Donath, T.

    2007-01-01

    Synchrotron-radiation-based computed microtomography (SR μ CT) is an established method for the examination of volume structures. It allows to measure the x-ray attenuation coefficient of a specimen three-dimensionally with a spatial resolution of about one micrometer. In contrast to conventional x-ray sources (x-ray tubes), the unique properties of synchrotron radiation enable quantitative measurements that do not suffer from beam-hardening artifacts. During this work the capabilities for quantitative SR μ CT measurements have been further improved by enhancements that were made to the SR μ CT apparatus and to the reconstruction chain. For high-resolution SR μ CT an x-ray camera consisting of luminescent screen (x-ray phosphor), lens system, and CCD camera was used. A significant suppression of blur that is caused by reflections inside the luminescent screen could be achieved by application of an absorbing optical coating to the screen surface. It is shown that blur and ring artifacts in the tomographic reconstructions are thereby drastically reduced. Furthermore, a robust and objective method for the determination of the center of rotation in projection data (sinograms) is presented that achieves sub-pixel precision. By implementation of this method into the reconstruction chain, complete automation of the reconstruction process has been achieved. Examples of quantitative SR μ CT studies conducted at the Hamburger Synchrotronstrahlungslabor HASYLAB at the Deutsches Elektronen-Synchrotron DESY are presented and used for the demonstration of the achieved enhancements. (orig.)

  1. Use of X-ray CT-defined regions of interest for the determination of SPECT recovery coefficients

    International Nuclear Information System (INIS)

    Tang, H.R.; Brown, J.K.; Hasegawa, B.H.

    1996-01-01

    For accurate activity per unit volume measurements in SPECT, recovery coefficients are usually applied based on the size and shape of objects being imaged to properly account for the resolution limitations of the gamma camera. Because of noise and limited spatial resolution, determination of object sizes and boundaries can be difficult using the SPECT images alone. We therefore have developed a technique which determines activity concentrations for SPECT using regions of interest (ROI's) obtained from coregistered X-ray CT images. In this study, experimental phantoms containing cylindrical and spherical objects were imaged on a combined X-ray CT/SPECT system and reconstructed data volumes were registered using the known geometry of the system. ROI's were defined on the registered CT images and used to help quantify activity concentration in localized regions and to measure object volumes. We have derived the recovery curves for these objects and scan technique. We have also tested a technique that demonstrates activity quantitation without the need for object and size dependent recovery coefficients in the case of low background

  2. X-Ray Psoralen Activated Cancer Therapy (X-PACT.

    Directory of Open Access Journals (Sweden)

    Mark Oldham

    Full Text Available This work investigates X-PACT (X-ray Psoralen Activated Cancer Therapy: a new approach for the treatment of solid cancer. X-PACT utilizes psoralen, a potent anti-cancer therapeutic with current application to proliferative disease and extracorporeal photopheresis (ECP of cutaneous T Cell Lymphoma. An immunogenic role for light-activated psoralen has been reported, contributing to long-term clinical responses. Psoralen therapies have to-date been limited to superficial or extracorporeal scenarios due to the requirement for psoralen activation by UVA light, which has limited penetration in tissue. X-PACT solves this challenge by activating psoralen with UV light emitted from novel non-tethered phosphors (co-incubated with psoralen that absorb x-rays and re-radiate (phosphoresce at UV wavelengths. The efficacy of X-PACT was evaluated in both in-vitro and in-vivo settings. In-vitro studies utilized breast (4T1, glioma (CT2A and sarcoma (KP-B cell lines. Cells were exposed to X-PACT treatments where the concentrations of drug (psoralen and phosphor and radiation parameters (energy, dose, and dose rate were varied. Efficacy was evaluated primarily using flow cell cytometry in combination with complimentary assays, and the in-vivo mouse study. In an in-vitro study, we show that X-PACT induces significant tumor cell apoptosis and cytotoxicity, unlike psoralen or phosphor alone (p<0.0001. We also show that apoptosis increases as doses of phosphor, psoralen, or radiation increase. Finally, in an in-vivo pilot study of BALBc mice with syngeneic 4T1 tumors, we show that the rate of tumor growth is slower with X-PACT than with saline or AMT + X-ray (p<0.0001. Overall these studies demonstrate a potential therapeutic effect for X-PACT, and provide a foundation and rationale for future studies. In summary, X-PACT represents a novel treatment approach in which well-tolerated low doses of x-ray radiation are delivered to a specific tumor site to generate UVA

  3. Structural changes of green roof growing substrate layer studied by X-ray CT

    Science.gov (United States)

    Jelinkova, Vladimira; Sacha, Jan; Dohnal, Michal; Snehota, Michal

    2017-04-01

    Increasing interest in green infrastructure linked with newly implemented legislation/rules/laws worldwide opens up research potential for field of soil hydrology. A better understanding of function of engineered soils involved in green infrastructure solutions such as green roofs or rain garden is needed. A soil layer is considered as a highly significant component of the aforesaid systems. In comparison with a natural soil, the engineered soil is assumed to be the more challenging case due to rapid structure changes early stages after its build-up. The green infrastructure efficiency depends on the physical and chemical properties of the soil, which are, in the case of engineered soils, a function of its initial composition and subsequent soil formation processes. The project presented in this paper is focused on fundamental processes in the relatively thick layer of engineered soil. The initial structure development, during which the pore geometry is altered by the growth of plant roots, water influx, solid particles translocation and other soil formation processes, is investigated with the help of noninvasive imaging technique  X-ray computed tomography. The soil development has been studied on undisturbed soil samples taken periodically from green roof test system during early stages of its life cycle. Two approaches and sample sizes were employed. In the first approach, undisturbed samples (volume of about 63 cm3) were taken each time from the test site and scanned by X-ray CT. In the second approach, samples (volume of about 630 cm3) were permanently installed at the test site and has been repeatedly removed to perform X-ray CT imaging. CT-derived macroporosity profiles reveal significant temporal changes of soil structure. Clogging of pores by fine particles and fissures development are two most significant changes that would affect the green roof system efficiency. This work has been supported by the Ministry of Education, Youth and Sports within

  4. Three dimensional reconstruction of fossils with X-ray CT and computer graphics

    Energy Technology Data Exchange (ETDEWEB)

    Hamada, Takashi; Tateno, Satoko (Tokyo Univ. (Japan). Coll. of Arts and Sciences); Suzuki, Naoki

    1991-12-01

    We have developed a method for three dimensional (3D) visualization of fossils such as trilobites and ammonites by non-destructive measurement and computer graphics. The imaging techniques in the medical sciences are applied for fossils by us to have quantitative data analyses on the structural and functional features of some extinct creatures. These methods are composed of a high resolutional X-ray computed tomography (X-ray CT) and computer graphics. We are able to observe not only outer shape but also inner structure of fossils as a 3D image by this method. Consequently, the shape and volume are measurable on these 3D image quantitatively. In addition to that, it is able to reconstruct an ideal figure from the deformed fossils by graphical treatments of the data. Such a 3D reconstruction method is useful to obtain a new information from the paleontological standpoint. (author).

  5. Three dimensional reconstruction of fossils with X-ray CT and computer graphics

    International Nuclear Information System (INIS)

    Hamada, Takashi; Tateno, Satoko; Suzuki, Naoki.

    1991-01-01

    We have developed a method for three dimensional (3D) visualization of fossils such as trilobites and ammonites by non-destructive measurement and computer graphics. The imaging techniques in the medical sciences are applied for fossils by us to have quantitative data analyses on the structural and functional features of some extinct creatures. These methods are composed of a high resolutional X-ray computed tomography (X-ray CT) and computer graphics. We are able to observe not only outer shape but also inner structure of fossils as a 3D image by this method. Consequently, the shape and volume are measurable on these 3D image quantitatively. In addition to that, it is able to reconstruct an ideal figure from the deformed fossils by graphical treatments of the data. Such a 3D reconstruction method is useful to obtain a new information from the paleontological standpoint. (author)

  6. Early Detection of Amyloid Plaque in Alzheimer’s Disease via X-ray Phase CT

    Science.gov (United States)

    2016-08-01

    Geraghty, J. A. Seibert, and S. L. Wootton-Gorges, “Dose reduction in pediatric CT : A rational approach,” Radiology 228, 352–360 (2003). Medical...improvement in image quality can be quantified as large as 450 [Fig. 6(c)] and 350 [Fig. 6(d)] Hounsfield units (HUs), where the Hounsfield unit is defined...AWARD NUMBER: W81XWH-12-1-0138 TITLE: Early Detection of Amyloid Plaque in Alzheimer’s Disease via X-ray Phase CT PRINCIPAL INVESTIGATOR

  7. Trabecular bone analysis in CT and X-ray images of the proximal femur for the assessment of local bone quality.

    Science.gov (United States)

    Fritscher, Karl; Grunerbl, Agnes; Hanni, Markus; Suhm, Norbert; Hengg, Clemens; Schubert, Rainer

    2009-10-01

    Currently, conventional X-ray and CT images as well as invasive methods performed during the surgical intervention are used to judge the local quality of a fractured proximal femur. However, these approaches are either dependent on the surgeon's experience or cannot assist diagnostic and planning tasks preoperatively. Therefore, in this work a method for the individual analysis of local bone quality in the proximal femur based on model-based analysis of CT- and X-ray images of femur specimen will be proposed. A combined representation of shape and spatial intensity distribution of an object and different statistical approaches for dimensionality reduction are used to create a statistical appearance model in order to assess the local bone quality in CT and X-ray images. The developed algorithms are tested and evaluated on 28 femur specimen. It will be shown that the tools and algorithms presented herein are highly adequate to automatically and objectively predict bone mineral density values as well as a biomechanical parameter of the bone that can be measured intraoperatively.

  8. Radiation dose reduction in medical x-ray CT via Fourier-based iterative reconstruction.

    Science.gov (United States)

    Fahimian, Benjamin P; Zhao, Yunzhe; Huang, Zhifeng; Fung, Russell; Mao, Yu; Zhu, Chun; Khatonabadi, Maryam; DeMarco, John J; Osher, Stanley J; McNitt-Gray, Michael F; Miao, Jianwei

    2013-03-01

    A Fourier-based iterative reconstruction technique, termed Equally Sloped Tomography (EST), is developed in conjunction with advanced mathematical regularization to investigate radiation dose reduction in x-ray CT. The method is experimentally implemented on fan-beam CT and evaluated as a function of imaging dose on a series of image quality phantoms and anonymous pediatric patient data sets. Numerical simulation experiments are also performed to explore the extension of EST to helical cone-beam geometry. EST is a Fourier based iterative algorithm, which iterates back and forth between real and Fourier space utilizing the algebraically exact pseudopolar fast Fourier transform (PPFFT). In each iteration, physical constraints and mathematical regularization are applied in real space, while the measured data are enforced in Fourier space. The algorithm is automatically terminated when a proposed termination criterion is met. Experimentally, fan-beam projections were acquired by the Siemens z-flying focal spot technology, and subsequently interleaved and rebinned to a pseudopolar grid. Image quality phantoms were scanned at systematically varied mAs settings, reconstructed by EST and conventional reconstruction methods such as filtered back projection (FBP), and quantified using metrics including resolution, signal-to-noise ratios (SNRs), and contrast-to-noise ratios (CNRs). Pediatric data sets were reconstructed at their original acquisition settings and additionally simulated to lower dose settings for comparison and evaluation of the potential for radiation dose reduction. Numerical experiments were conducted to quantify EST and other iterative methods in terms of image quality and computation time. The extension of EST to helical cone-beam CT was implemented by using the advanced single-slice rebinning (ASSR) method. Based on the phantom and pediatric patient fan-beam CT data, it is demonstrated that EST reconstructions with the lowest scanner flux setting of 39 m

  9. Radiation dose reduction in medical x-ray CT via Fourier-based iterative reconstruction

    International Nuclear Information System (INIS)

    Fahimian, Benjamin P.; Zhao Yunzhe; Huang Zhifeng; Fung, Russell; Zhu Chun; Miao Jianwei; Mao Yu; Khatonabadi, Maryam; DeMarco, John J.; McNitt-Gray, Michael F.; Osher, Stanley J.

    2013-01-01

    Purpose: A Fourier-based iterative reconstruction technique, termed Equally Sloped Tomography (EST), is developed in conjunction with advanced mathematical regularization to investigate radiation dose reduction in x-ray CT. The method is experimentally implemented on fan-beam CT and evaluated as a function of imaging dose on a series of image quality phantoms and anonymous pediatric patient data sets. Numerical simulation experiments are also performed to explore the extension of EST to helical cone-beam geometry. Methods: EST is a Fourier based iterative algorithm, which iterates back and forth between real and Fourier space utilizing the algebraically exact pseudopolar fast Fourier transform (PPFFT). In each iteration, physical constraints and mathematical regularization are applied in real space, while the measured data are enforced in Fourier space. The algorithm is automatically terminated when a proposed termination criterion is met. Experimentally, fan-beam projections were acquired by the Siemens z-flying focal spot technology, and subsequently interleaved and rebinned to a pseudopolar grid. Image quality phantoms were scanned at systematically varied mAs settings, reconstructed by EST and conventional reconstruction methods such as filtered back projection (FBP), and quantified using metrics including resolution, signal-to-noise ratios (SNRs), and contrast-to-noise ratios (CNRs). Pediatric data sets were reconstructed at their original acquisition settings and additionally simulated to lower dose settings for comparison and evaluation of the potential for radiation dose reduction. Numerical experiments were conducted to quantify EST and other iterative methods in terms of image quality and computation time. The extension of EST to helical cone-beam CT was implemented by using the advanced single-slice rebinning (ASSR) method. Results: Based on the phantom and pediatric patient fan-beam CT data, it is demonstrated that EST reconstructions with the lowest

  10. CT breast dose reduction with the use of breast positioning and organ-based tube current modulation.

    Science.gov (United States)

    Fu, Wanyi; Tian, Xiaoyu; Sturgeon, Gregory M; Agasthya, Greeshma; Segars, William Paul; Goodsitt, Mitchell M; Kazerooni, Ella A; Samei, Ehsan

    2017-02-01

    This study aimed to investigate the breast dose reduction potential of a breast-positioning (BP) technique for thoracic CT examinations with organ-based tube current modulation (OTCM). This study included 13 female anthropomorphic computational phantoms (XCAT, age range: 27-65 y.o., weight range: 52-105.8 kg). Each phantom was modified to simulate three breast sizes in standard supine geometry. The modeled breasts were then morphed to emulate BP that constrained the majority of the breast tissue inside the 120° anterior tube current (mA) reduction zone. The OTCM mA value was modeled using a ray-tracing program, which reduced the mA to 20% in the anterior region with a corresponding increase to the posterior region. The organ doses were estimated by a validated Monte Carlo program for a typical clinical CT system (SOMATOM Definition Flash, Siemens Healthcare). The simulated organ doses and organ doses normalized by CTDI vol were used to compare three CT protocols: attenuation-based tube current modulation (ATCM), OTCM, and OTCM with BP (OTCM BP ). On average, compared to ATCM, OTCM reduced breast dose by 19.3 ± 4.5%, whereas OTCM BP reduced breast dose by 38.6 ± 8.1% (an additional 23.8 ± 9.4%). The dose saving of OTCM BP was more significant for larger breasts (on average 33, 38, and 44% reduction for 0.5, 1, and 2 kg breasts, respectively). Compared to ATCM, OTCM BP also reduced thymus and heart dose by 15.1 ± 7.4% and 15.9 ± 6.2% respectively. In thoracic CT examinations, OTCM with a breast-positioning technique can markedly reduce unnecessary exposure to radiosensitive organs in anterior chest wall, specifically breast tissue. The breast dose reduction is more notable for women with larger breasts. © 2016 American Association of Physicists in Medicine.

  11. Applications of phase-contrast x-ray imaging to medicine using an x-ray interferometer

    Science.gov (United States)

    Momose, Atsushi; Yoneyama, Akio; Takeda, Tohoru; Itai, Yuji; Tu, Jinhong; Hirano, Keiichi

    1999-10-01

    We are investigating possible medical applications of phase- contrast X-ray imaging using an X-ray interferometer. This paper introduces the strategy of the research project and the present status. The main subject is to broaden the observation area to enable in vivo observation. For this purpose, large X-ray interferometers were developed, and 2.5 cm X 1.5 cm interference patterns were generated using synchrotron X-rays. An improvement of the spatial resolution is also included in the project, and an X-ray interferometer designed for high-resolution phase-contrast X-ray imaging was fabricated and tested. In parallel with the instrumental developments, various soft tissues are observed by phase- contrast X-ray CT to find correspondence between the generated contrast and our histological knowledge. The observation done so far suggests that cancerous tissues are differentiated from normal tissues and that blood can produce phase contrast. Furthermore, this project includes exploring materials that modulate phase contrast for selective imaging.

  12. Body packers on your examination table: How helpful are plain x-ray images? A definitive low-dose CT protocol as a diagnosis tool for body packers.

    Science.gov (United States)

    Schulz, B; Grossbach, A; Gruber-Rouh, T; Zangos, S; Vogl, Th J; Eichler, K

    2014-12-01

    To analyze the clinical value and radiation dose of plain x-rays and CT in examining patients suspected of ingesting drug-filled packets. Thirty-eight patients with suspected internal concealment of drug-filled packets who were examined with plain x-rays or CT or both were included in the study. CT studies were performed using low-dose and standard-dose techniques. All radiographic images were analysed by two radiologists regarding identification of the packets and estimating the effective radiation dose from standard- and low-dose CT versus conventional x-ray examinations. Descriptive calculations were made regarding the number and density of packs and radiation dosage. The diagnostic performance of both radiologists with standard- and low-dose CT was calculated by analysing differences in the mean number of packs found. Thirty-one patients were positively identified as body packers with an average of 13 packs (min: n = 1, max: n = 58, total: n = 390); seven patients were not concealing drug packets. X-ray images were taken of 24 patients prior to CT, thus allowing a direct comparison between the two methods. The correct diagnosis was made in 42%, in 33% the radiologists were uncertain, and in 25% of drug packets were either not or wrongly identified. X-ray imaging had a positive predictive value of 20% with a negative predictive value of 81%. A total of 55 CT examinations were performed on all patients with a mean effective dose of 2 mSv (low dose) versus 9.3 mSv (standard dose). The visibility of packets on low-dose CT images compared to high-dose CT was not reduced: the radiologists identified 385 and 381 of the packets, respectively, with no difference regarding the examination technique (p = 0.24 and p = 0.253, respectively). The radiodensity of all drug-filled packets at CT ranged from 26-292 HU (mean 181.2 HU). X-ray imaging of supposed body packers leads to a significant risk of diagnostic errors and additional need for CT. Instead, a single abdominal low

  13. Detection of breast abnormalities on enhanced chest CT: Correlation with breast composition on mammography

    Energy Technology Data Exchange (ETDEWEB)

    Cho, Eun Mi; Kang, Hee; Shin, Young Gyung; Yun, Jong Hyouk; Oh, Kyung Seung [Dept. of Radiology, Kosin University Gospel Hospital, Kosin University College of Medicine, Busan (Korea, Republic of)

    2017-02-15

    To investigate the capability of enhanced chest computed tomography (CT) for detecting breast abnormalities and to assess the influence of breast composition on this detectability. From 2000 to 2013, 75 patients who underwent mammography, breast sonography, and enhanced chest CT within one month and had abnormalities on sonography were included. Detection rate of breast abnormality on enhanced chest CT was compared among 4 types of breast composition by the Breast Imaging Reporting and Data System. Contribution of breast composition, size and enhancement of target lesions to detectability of enhanced chest CT was assessed using logistic regression and chi-square test. Of the 75 target lesions, 34 (45.3%) were detected on enhanced chest CT, corresponding with those on breast sonography; there were no significantly different detection rates among the 4 types of breast composition (p = 0.078). Breast composition [odds ratio (OR) = 1.07, p = 0.206] and enhancement (OR = 21.49, p = 0.998) had no significant effect, but size (OR = 1.23, p = 0.004) was a significant contributing factor influencing the detectability of enhanced chest CT for breast lesions. About half of the cases (45.3%) demonstrated breast lesions on chest CT corresponding with target lesions on sonography. Breast composition defined on mammography did not affect the detectability of enhanced chest CT for breast lesions.

  14. Cone-beam micro computed tomography dedicated to the breast.

    Science.gov (United States)

    Sarno, Antonio; Mettivier, Giovanni; Di Lillo, Francesca; Cesarelli, Mario; Bifulco, Paolo; Russo, Paolo

    2016-12-01

    We developed a scanner for micro computed tomography dedicated to the breast (BµCT) with a high resolution flat-panel detector and a microfocus X-ray tube. We evaluated the system spatial resolution via the 3D modulation transfer function (MTF). In addition to conventional absorption-based X-ray imaging, such a prototype showed capabilities for propagation-based phase-contrast and related edge enhancement effects in 3D imaging. The system limiting spatial resolution is 6.2mm -1 (MTF at 10%) in the vertical direction and 3.8mm -1 in the radial direction, values which compare favorably with the spatial resolution reached by mini focus breast CT scanners of other groups. The BµCT scanner was able to detect both microcalcification clusters and masses in an anthropomorphic breast phantom at a dose comparable to that of two-view mammography. The use of a breast holder is proposed in order to have 1-2min long scan times without breast motion artifacts. Copyright © 2016 IPEM. Published by Elsevier Ltd. All rights reserved.

  15. Identification of human breast pathologies by X-ray elastic scattering; Identificacao de patologias mamarias atraves do espalhamento elastico de raios X

    Energy Technology Data Exchange (ETDEWEB)

    Conceicao, Andre L.C.; Antoniassi, Marcelo; Poletti, Martin E., E-mail: andre_conceicao@yahoo.com.b [Universidade de Sao Paulo (FFCLRP/USP), Ribeirao Preto, SP (Brazil). Faculdade de Filosofia, Ciencias e Letras. Dept. de Fisica e Matematica

    2011-07-01

    In this paper we determine the scattering profiles of normal, benign and malignant human breast samples in a momentum transfer range of 0.07nm{sup -1} {<=}q{<=}70.55nm{sup -1}, resulted from combining WAXS (wide angle x-ray scattering) and SAXS (small angle x-ray scattering) data. The results showed considerable differences between the scattering profiles of each tissue type. Based on this fact, some parameters, representing structural features, were extracted from these scattering profiles and submitted to a discriminant analysis. From statistical analysis, the ratio between the peak intensities at q=19.8nm{sup -1} and q=13.9nm{sup -1} and the intensity of third order axial collagen peak arose as two potentials breast tissue classifiers and, from combining them it was possible differentiate among normal, benign and malignant lesions. (author)

  16. Engineering iodine-doped carbon dots as dual-modal probes for fluorescence and X-ray CT imaging

    Directory of Open Access Journals (Sweden)

    Zhang M

    2015-11-01

    Full Text Available Miaomiao Zhang,1,* Huixiang Ju,2,* Li Zhang,1,* Mingzhong Sun,2 Zhongwei Zhou,2 Zhenyu Dai,3 Lirong Zhang,1 Aihua Gong,1 Chaoyao Wu,1 Fengyi Du1 1School of Medicine, Jiangsu University, Zhenjiang, People’s Republic of China; 2Department of Clinical Laboratory, Affiliated Yancheng Hospital, School of Medicine, Southeast University, Yancheng, Jiangsu, People’s Republic of China; 3Radiology Department, Affiliated Yancheng Hospital, School of Medicine, Southeast University, Yancheng, Jiangsu, People’s Republic of China *These authors contributed equally to this work Abstract: X-ray computed tomography (CT is the most commonly used imaging technique for noninvasive diagnosis of disease. In order to improve tissue specificity and prevent adverse effects, we report the design and synthesis of iodine-doped carbon dots (I-doped CDs as efficient CT contrast agents and fluorescence probe by a facile bottom-up hydrothermal carbonization process. The as-prepared I-doped CDs are monodispersed spherical nanoparticles (a diameter of ~2.7 nm with favorable dispersibility and colloidal stability in water. The aqueous solution of I-doped CDs showed wavelength-dependent excitation and stable photoluminescence similar to traditional carbon quantum dots. Importantly, I-doped CDs displayed superior X-ray attenuation properties in vitro and excellent biocompatibility. After intravenous injection, I-doped CDs were distributed throughout the body and excreted by renal clearance. These findings validated that I-doped CDs with high X-ray attenuation potency and favorable photoluminescence show great promise for biomedical research and disease diagnosis. Keywords: carbon dots, contrast agents, iodine-doped, CT imaging

  17. Quantitative X-ray CT analysis of calcification of the abdominal aorta and its relationship to obesity

    International Nuclear Information System (INIS)

    Shinagawa, Toshio; Hiraiwa, Yoshio; Mizuno, Seio; Kusunoki, Norio; Nitta, Yu; Matsubara, Takao; Iwainaka, Yoichi; Konishi, Hideo

    1992-01-01

    Quantitative analysis of abdominal aorta calcification by X-ray CT is useful method for non-invasive diagnosis of atherosclerosis. We recently examined the relationship between the X-ray CT measurement of abdominal aorta calcification and the degree of obesity. For this purpose, the body mass index (BMI) and the subcutaneous fat thickness (determined by X-ray CT at the umbilical level) were analyzed in relation to the abdominal aorta calcification index (ACI) in 845 patients (453 males and 392 females aged 40-79 years). Patients with BMI under 20 were classified as 'lean', those with BMI between 20-26 as 'normal' and those with BMI over 26 as 'obese'. 1. Among males, the ACI was highest in lean individuals and lowest in obese individuals. The difference in ACI between lean and obese males was significant in the middle aged group (40-65 years). Among females, no relationship was observed between the degree of obesity and ACI. 2. Among males, ACI was higher in individuals with low subcutaneous fat thickness and lower in individuals with greater subcutaneous fat thickness. The difference was significant in the middle aged group. Among females, no relationship was observed between the two parameters. 3. When the visceral fat to subcutaneous fat ratio (V/S) in 85 males and females aged 60-69 years was analyzed in relation to ACI, ACI tended to decrease as the V/S increased, in both males and females. 4. Relationships between BMI and subcutaneous fat thickness, between BMI and lipids and between lipids and ACI were also analyzed. (author)

  18. Breast density quantification with cone-beam CT: a post-mortem study

    International Nuclear Information System (INIS)

    Johnson, Travis; Ding, Huanjun; Le, Huy Q; Ducote, Justin L; Molloi, Sabee

    2013-01-01

    Forty post-mortem breasts were imaged with a flat-panel based cone-beam x-ray CT system at 50 kVp. The feasibility of breast density quantification has been investigated using standard histogram thresholding and an automatic segmentation method based on the fuzzy c-means algorithm (FCM). The breasts were chemically decomposed into water, lipid, and protein immediately after image acquisition was completed. The per cent fibroglandular volume (%FGV) from chemical analysis was used as the gold standard for breast density comparison. Both image-based segmentation techniques showed good precision in breast density quantification with high linear coefficients between the right and left breast of each pair. When comparing with the gold standard using %FGV from chemical analysis, Pearson's r-values were estimated to be 0.983 and 0.968 for the FCM clustering and the histogram thresholding techniques, respectively. The standard error of the estimate was also reduced from 3.92% to 2.45% by applying the automatic clustering technique. The results of the postmortem study suggested that breast tissue can be characterized in terms of water, lipid and protein contents with high accuracy by using chemical analysis, which offers a gold standard for breast density studies comparing different techniques. In the investigated image segmentation techniques, the FCM algorithm had high precision and accuracy in breast density quantification. In comparison to conventional histogram thresholding, it was more efficient and reduced inter-observer variation. (paper)

  19. Sparse signal reconstruction from polychromatic X-ray CT measurements via mass attenuation discretization

    International Nuclear Information System (INIS)

    Gu, Renliang; Dogandžić, Aleksandar

    2014-01-01

    We propose a method for reconstructing sparse images from polychromatic x-ray computed tomography (ct) measurements via mass attenuation coefficient discretization. The material of the inspected object and the incident spectrum are assumed to be unknown. We rewrite the Lambert-Beer’s law in terms of integral expressions of mass attenuation and discretize the resulting integrals. We then present a penalized constrained least-squares optimization approach for reconstructing the underlying object from log-domain measurements, where an active set approach is employed to estimate incident energy density parameters and the nonnegativity and sparsity of the image density map are imposed using negative-energy and smooth ℓ 1 -norm penalty terms. We propose a two-step scheme for refining the mass attenuation discretization grid by using higher sampling rate over the range with higher photon energy, and eliminating the discretization points that have little effect on accuracy of the forward projection model. This refinement allows us to successfully handle the characteristic lines (Dirac impulses) in the incident energy density spectrum. We compare the proposed method with the standard filtered backprojection, which ignores the polychromatic nature of the measurements and sparsity of the image density map. Numerical simulations using both realistic simulated and real x-ray ct data are presented

  20. Body packers on your examination table: How helpful are plain x-ray images? A definitive low-dose CT protocol as a diagnosis tool for body packers

    International Nuclear Information System (INIS)

    Schulz, B.; Grossbach, A.; Gruber-Rouh, T.; Zangos, S.; Vogl, Th. J.; Eichler, K.

    2014-01-01

    Aim: To analyze the clinical value and radiation dose of plain x-rays and CT in examining patients suspected of ingesting drug-filled packets. Materials and methods: Thirty-eight patients with suspected internal concealment of drug-filled packets who were examined with plain x-rays or CT or both were included in the study. CT studies were performed using low-dose and standard-dose techniques. All radiographic images were analysed by two radiologists regarding identification of the packets and estimating the effective radiation dose from standard- and low-dose CT versus conventional x-ray examinations. Descriptive calculations were made regarding the number and density of packs and radiation dosage. The diagnostic performance of both radiologists with standard- and low-dose CT was calculated by analysing differences in the mean number of packs found. Results: Thirty-one patients were positively identified as body packers with an average of 13 packs (min: n = 1, max: n = 58, total: n = 390); seven patients were not concealing drug packets. X-ray images were taken of 24 patients prior to CT, thus allowing a direct comparison between the two methods. The correct diagnosis was made in 42%, in 33% the radiologists were uncertain, and in 25% of drug packets were either not or wrongly identified. X-ray imaging had a positive predictive value of 20% with a negative predictive value of 81%. A total of 55 CT examinations were performed on all patients with a mean effective dose of 2 mSv (low dose) versus 9.3 mSv (standard dose). The visibility of packets on low-dose CT images compared to high-dose CT was not reduced: the radiologists identified 385 and 381 of the packets, respectively, with no difference regarding the examination technique (p = 0.24 and p = 0.253, respectively). The radiodensity of all drug-filled packets at CT ranged from 26–292 HU (mean 181.2 HU). Conclusion: X-ray imaging of supposed body packers leads to a significant risk of diagnostic

  1. Comparison of the efficacy of preoperative X-ray and thermoradiotherapy used for the treatment of breast cancer

    International Nuclear Information System (INIS)

    Muravskaya, G.V.; Pantyushenko, T.A.; Fradkin, S.Z.; Zhavrid, Eh.A.; Moiseenko, V.V.

    1984-01-01

    An experience of the first randomatized clinical investigation on the usage of different variants of preoperative X-ray (with simultaneous local UHF hyperthepmia and without it) action in the case of combined treatment of patients with breast cancer has been summarized. It has been shown that conventional values of preoperative X-ray therapy (about 30-45 G.) are optimum ones. An increase of complex thermoradiotherapy efficacy in the case of considered cancer forms may be achieved at the expense of a stre gthening of preoperative X-ray action by means of dose increase up to the cancerogenic level or preoperative irradiation under the conditions of local UHF-hyperthermia

  2. Automated analysis for early signs of cerebral infarctions on brain X-ray CT images

    International Nuclear Information System (INIS)

    Oshima, Kazuki; Hara, Takeshi; Zhou, X.; Muramatsu, Chisako; Fujita, Hiroshi; Sakashita, Keiji

    2010-01-01

    t-PA (tissue plasminogen activator) thrombolysis is an effective clinical treatment for the acute cerebral infarction by breakdown to blood clots. However there is a risk of hemorrhage with its use. The guideline of the treatment is denying cerebral hemorrhage and widespread Early CT sign (ECS) on CT images. In this study, we analyzed the CT value of normal brain and ECS with normal brain model by comparing patient brain CT scan with a statistical normal model. Our method has constructed normal brain models consisted of 60 normal brain X-ray CT images. We calculated Z-score based on statistical model for 16 cases of cerebral infarction with ECS, 3 cases of cerebral infarction without ECS, and 25 cases of normal brain. The results of statistical analysis showed that there was a statistically significant difference between control and abnormal groups. This result implied that the automated detection scheme for ECS by using Z-score would be a possible application for brain computer-aided diagnosis (CAD). (author)

  3. X-ray CT analysis on archaeological iron based artifacts

    International Nuclear Information System (INIS)

    Honda, Takashi; Gunji, Eiichi

    2005-01-01

    Corrosion analysis was carried out on twenty iron-plates, which had been dug out at the 6th Yamato ancient tomb (Nara-shi). It was evaluated through the inner-structural analysis by X-ray CT and the XRD and chemical analysis of the rusts that the most iron-plates had been buried in a slightly oxidizing environment and the maximum corrosion depth was 1.6 mm for about 1500 years. The analysis result of the extreme-value data indicates that the maximum depth of an overpack is estimated to be 2.5-2.7 mm. A part of small iron-plates were supposed to have been buried in an oxidizing environment, and about 1 mm-thick rust layers with cavities existed and the corrosion amount was figured out to be about 0.3 mm. (author)

  4. Correlation analysis between expression of PCNA, Ki-67 and COX-2 and X-ray features in mammography in breast cancer.

    Science.gov (United States)

    Qiu, Xiaoming; Mei, Jixin; Yin, Jianjun; Wang, Hong; Wang, Jinqi; Xie, Ming

    2017-09-01

    This study investigated expression of proliferating cell nuclear antigen (PCNA), proliferation-associated nuclear antigen (Ki-67) and cyclooxygenase-2 (COX-2) in tissues of breast invasive ductal carcinoma, and analyzed the correlations between these indexes and X-ray features in mammography. A total of 90 patients who were admitted to Huangshi Central Hospital and diagnosed as breast invasive ductal carcinoma from January 2014 to January 2016 were selected. The expression of PCNA, Ki-67 and COX-2 in cancer tissues and cancer-adjacent normal tissues of patients were detected by immunohistochemical staining, and X-ray features in mammography of patients were observed. By using Spearman correlation analysis, the correlations between expression of PCNA, Ki-67 and COX-2 and X-ray features in mammography in breast cancer were investigated. As a result, the positive expression rates of PCNA, Ki-67 and COX-2 in cancer tissues of the patient groups were respectively 42.2, 45.6 and 51.1%, which were significantly higher than those in cancer-adjacent normal tissues of the control group (pcorrelation with age and tumor size (p>0.05). PCNA, Ki-67 and COX-2 expression in cancer tissues of the patient group had no correlation with the existence of lumps and localized density-increased shadows (p>0.05), but were associated with manifestations of architectural distortion, calcification as well as skin and nipple depression (pcorrelation analysis revealed that there was a significantly positive correlation between the expression of PCNA and COX-2 in cancer tissues of the patient group (r=0.676, pcorrelation between the expression of Ki-67 and COX-2 (r=0.724, pcorrelation with the expression of Ki-67 (p>0.05). In conclusion, PCNA, Ki-67 and COX-2 expression is of great significance in the occurrence, invasion and metastasis of breast invasive ductal carcinoma. There is a strong correlation between PCNA, Ki-67 and COX-2 expression levels and X-ray features in mammography in breast

  5. Synchrotron x-ray studies of the keel of the short-spined sea urchin lytechinus variegatus: absorption microtomography (microCT) and small beam diffraction mapping

    International Nuclear Information System (INIS)

    Stock, S.R.; Barss, J.; Dahl, T.; Veis, A.; Almer, J.D.; De Carlo, F.

    2003-01-01

    In sea urchin teeth, the keel plays an important structural role, and this paper reports results of microstructural characterization of the keel of Lytechinus variegatus using two noninvasive synchrotron x-ray techniques: x-ray absorption microtomography (microCT) and x-ray diffraction mapping. MicroCT with 14 keV x-rays mapped the spatial distribution of mineral at the 1.3 microm level in a millimeter-sized fragment of a mature portion of the keel. Two rows of low absorption channels (i.e., primary channels) slightly less than 10 microm in diameter were found running linearly from the flange to the base of the keel and parallel to its sides. The primary channels paralleled the oral edge of the keel, and the microCT slices revealed a planar secondary channel leading from each primary channel to the side of the keel. The primary and secondary channels were more or less coplanar and may correspond to the soft tissue between plates of the carinar process. Transmission x-ray diffraction with 80.8 keV x-rays and a 0.1 mm beam mapped the distribution of calcite crystal orientations and the composition Ca(1-x)Mg(x)CO(3) of the calcite. Unlike the variable Mg concentration and highly curved prisms found in the keel of Paracentrotus lividus, a constant Mg content (x = 0.13) and relatively little prism curvature was found in the keel of Lytechinus variegatus.

  6. High resolution stationary digital breast tomosynthesis using distributed carbon nanotube x-ray source array.

    Science.gov (United States)

    Qian, Xin; Tucker, Andrew; Gidcumb, Emily; Shan, Jing; Yang, Guang; Calderon-Colon, Xiomara; Sultana, Shabana; Lu, Jianping; Zhou, Otto; Spronk, Derrek; Sprenger, Frank; Zhang, Yiheng; Kennedy, Don; Farbizio, Tom; Jing, Zhenxue

    2012-04-01

    The purpose of this study is to investigate the feasibility of increasing the system spatial resolution and scanning speed of Hologic Selenia Dimensions digital breast tomosynthesis (DBT) scanner by replacing the rotating mammography x-ray tube with a specially designed carbon nanotube (CNT) x-ray source array, which generates all the projection images needed for tomosynthesis reconstruction by electronically activating individual x-ray sources without any mechanical motion. The stationary digital breast tomosynthesis (s-DBT) design aims to (i) increase the system spatial resolution by eliminating image blurring due to x-ray tube motion and (ii) reduce the scanning time. Low spatial resolution and long scanning time are the two main technical limitations of current DBT technology. A CNT x-ray source array was designed and evaluated against a set of targeted system performance parameters. Simulations were performed to determine the maximum anode heat load at the desired focal spot size and to design the electron focusing optics. Field emission current from CNT cathode was measured for an extended period of time to determine the stable life time of CNT cathode for an expected clinical operation scenario. The source array was manufactured, tested, and integrated with a Selenia scanner. An electronic control unit was developed to interface the source array with the detection system and to scan and regulate x-ray beams. The performance of the s-DBT system was evaluated using physical phantoms. The spatially distributed CNT x-ray source array comprised 31 individually addressable x-ray sources covering a 30 angular span with 1 pitch and an isotropic focal spot size of 0.6 mm at full width at half-maximum. Stable operation at 28 kV(peak) anode voltage and 38 mA tube current was demonstrated with extended lifetime and good source-to-source consistency. For the standard imaging protocol of 15 views over 14, 100 mAs dose, and 2 × 2 detector binning, the projection

  7. Classification of fractures of the distal r[ius - Comparative evaluation of spiral CT images and X-rays

    International Nuclear Information System (INIS)

    Roehnert, W.; Nitzsche, H.; Franck, W.M.; Amlang, M.

    1998-01-01

    Within the framework of a prospective study 35 patients with fractures of the distal r[ius safely diagnosed to require surgery have been [ditionally scanned by spiral CT. For more exact diagnostic evaluation of the joint surfaces and the fractures, multiplane reconstructions have been m[e to enhance the information obtained from the primary, axial tomographic images. Two experts independently performed classification on the basis of the X-rays and the CT scans, applying the AO scheme and the method of Frykman. Classifications according to the AO scheme were found to agree for only 28.6 % of the patients, which means that 57.1 % of the fractures shown by the X-rays were more or less underassessed. Classification according to Frykman's method revealed fewer deviations and agreement in 51.4 % of the cases. Biplanar r[iography continues to be considered as the standard method for examination of distal r[ius fractures, but it is recommended to perform spiral CT scans for evaluation of more complex distal r[ius fractures because these may require different therapies according to fracture types. (orig./CB) [de

  8. X-ray CT high-density artefact suppression in the presence of bones

    Energy Technology Data Exchange (ETDEWEB)

    Wei Jikun [School of Health Sciences, 550 Stadium Mall Drive, West Lafayette, IN 47907-2051 (United States); Chen Laigao [BioImaging Center of Emphasis, Pfizer Global Research and Development, 2800 Plymouth RD, Ann Arbor, MI 48105 (United States); Sandison, George A [School of Health Sciences, 550 Stadium Mall Drive, West Lafayette, IN 47907-2051 (United States); Liang Yun [Department of Radiology, Indiana University Medical School, Indianapolis, IN 46202 (United States); Xu, Lisa X [School of Mechanical Engineering, 585 Purdue Mall, West Lafayette, IN 47907-2040 (United States)

    2004-12-21

    This paper presents a novel method of reducing x-ray CT high-density artefacts generated by metal objects when abundant bone structures are present in the region of interest. This method has an advantage over previously proposed methods since it heavily suppresses the metal artefacts without introducing extra bone artefacts. The method of suppression requires that bone pixels are isolated and segmented by thresholding. Then artificial CT numbers are assigned to the bone pixels so that their projection profiles are smooth and thus can be properly simulated by a polynomial interpolation. The projection profile of the metal object is then removed to fully suppress the artefacts. The resulting processed profile is fed to a reconstruction routine and the previously preserved bone pixels added back. The new method utilizes two important features of the CT image with metal artefacts: (a) metal and bone pixels are not severely affected by the high-density artefacts and (b) the high-density artefacts can be located in specific projection channels in the profile domain, although they are spread out in the image domain. This suppression method solves the problem of CT image artefacts arising from metal objects in the body. It has the potential to greatly improve diagnostic CT imaging in the presence of these objects and treatment planning that utilizes CT for patients with metal applicators (e.g., brachytherapy for cervix cancer and prostate cryotherapy)

  9. X-ray CT high-density artefact suppression in the presence of bones

    International Nuclear Information System (INIS)

    Wei Jikun; Chen Laigao; Sandison, George A; Liang Yun; Xu, Lisa X

    2004-01-01

    This paper presents a novel method of reducing x-ray CT high-density artefacts generated by metal objects when abundant bone structures are present in the region of interest. This method has an advantage over previously proposed methods since it heavily suppresses the metal artefacts without introducing extra bone artefacts. The method of suppression requires that bone pixels are isolated and segmented by thresholding. Then artificial CT numbers are assigned to the bone pixels so that their projection profiles are smooth and thus can be properly simulated by a polynomial interpolation. The projection profile of the metal object is then removed to fully suppress the artefacts. The resulting processed profile is fed to a reconstruction routine and the previously preserved bone pixels added back. The new method utilizes two important features of the CT image with metal artefacts: (a) metal and bone pixels are not severely affected by the high-density artefacts and (b) the high-density artefacts can be located in specific projection channels in the profile domain, although they are spread out in the image domain. This suppression method solves the problem of CT image artefacts arising from metal objects in the body. It has the potential to greatly improve diagnostic CT imaging in the presence of these objects and treatment planning that utilizes CT for patients with metal applicators (e.g., brachytherapy for cervix cancer and prostate cryotherapy)

  10. Introduction of a deformable x-ray CT polymer gel dosimetry system

    Science.gov (United States)

    Maynard, E.; Heath, E.; Hilts, M.; Jirasek, A.

    2018-04-01

    This study introduces the first 3D deformable dosimetry system based on x-ray computed tomography (CT) polymer gel dosimetry and establishes the setup reproducibility, deformation characteristics and dose response of the system. A N-isopropylacrylamide (NIPAM)-based gel formulation optimized for x-ray CT gel dosimetry was used, with a latex balloon serving as the deformable container and low-density polyethylene and polyvinyl alcohol providing additional oxygen barrier. Deformable gels were irradiated with a 6 MV calibration pattern to determine dosimetric response and a dosimetrically uniform plan to determine the spatial uniformity of the response. Wax beads were added to each gel as fiducial markers to track the deformation and setup of the gel dosimeters. From positions of the beads on CT images the setup reproducibility and the limits and reproducibility of gel deformation were determined. Comparison of gel measurements with Monte Carlo dose calculations found excellent dosimetric accuracy, comparable to that of an established non-deformable dosimetry system, with a mean dose discrepancy of 1.5% in the low-dose gradient region and a gamma pass rate of 97.9% using a 3%/3 mm criterion. The deformable dosimeter also showed good overall spatial dose uniformity throughout the dosimeter with some discrepancies within 20 mm of the edge of the container. Tracking of the beads within the dosimeter found that sub-millimetre setup accuracy is achievable with this system. The dosimeter was able to deform and relax when externally compressed by up to 30 mm without sustaining any permanent damage. Internal deformations in 3D produced average marker movements of up to 12 mm along the direction of compression. These deformations were also shown to be reproducible over 100 consecutive deformations. This work has established several important characteristics of a new deformable dosimetry system which shows promise for future clinical applications, including the

  11. A statistical survey of x-ray CT cases at the Clinic of Oral and Maxillofacial Radiology, Chiba Hospital, Tokyo Dental College

    International Nuclear Information System (INIS)

    Kitagawa, Hiromi; Wakoh, Mamoru; Shibuya, Hitoshi; Yamada, Masayuki; Harada, Takuya; Yamamoto, Kazuhiro; Makihara, Masahiro; Kuroyanagi, Kinya

    1997-01-01

    Statistical study was performed of x-ray CT cases at the Clinic of Oral and Maxillofacial Radiology, Chiba Hospital, Tokyo Dental College since a Toshiba CT Scanner TCT-700S was settled in 1988, until December 1994. Total number of cases photographed was 2645 cases, Male; 1447 (54.7%), Female; 1198 cases (45.3%). Total number of cases yearly photographed have increased every year. 95.43% of the cases were the diseases of oral surgery regions. X-ray CT has been used to malignant tumors (730 cases; 29.7%), cyst (435 cases; 17.7%), inflammation (362 cases; 14.7%), benign tumors (261 cases; 10.6%), injury (171 cases; 7.0%), salivary gland diseases (126 cases; 5.1%) and others. The number of tumors and cyst have been increasing every year. Average number of slices in every diseases were counted. Malignant tumor, injury, temporomandibular joint diseases, and congenital anomalies and malformations were counted many slices. Percentages of number of enhanced CT cases have been increased every year. Recently, number of enhanced CT cases have more number than non-enhanced CT cases. This attitude is correlated with the number of malignant tumors which have been increasing every year. Total number of cases of three dimensional imaging CT (3D-CT) was 316 cases. 3D-CT has been used to injury (146 cases; 46.2%), temporomandibular joint diseases (52 cases; 16.4%), congenital anomalies and malformations (49 cases; 15.5%), tumors (21 cases; 6.7%), cyst (13 cases; 4.1%) and others. The need of x-ray CT in our field and the tendency of dental treatment at Chiba Hospital might be changed in the future. In order to this situation, this type of statistical study will be performed again. (author)

  12. An experimental study on the noise properties of x-ray CT sinogram data in Radon space

    Energy Technology Data Exchange (ETDEWEB)

    Wang Jing; Liang Zhengrong; Eremina, Daria; Wang Su; Manzione, James [Department of Radiology, State University of New York, Stony Brook, NY 11794 (United States); Lu Hongbing [Department of Biomedical Engineering, Fourth Military Medical University, Xi' An, Shaanxi 710032 (China); Zhang Guangxiang; Chen, John [Department of Preventive Medicine, State University of New York, Stony Brook, NY 11794 (United States)], E-mail: jerome.liang@sunysb.edu

    2008-06-21

    Computed tomography (CT) has been well established as a diagnostic tool through hardware optimization and sophisticated data calibration. For screening purposes, the associated x-ray exposure risk must be minimized. An effective way to minimize the risk is to deliver fewer x-rays to the subject or lower the mAs parameter in data acquisition. This will increase the data noise. This work aims to study the noise property of the calibrated or preprocessed sinogram data in Radon space as the mAs level decreases. An anthropomorphic torso phantom was scanned repeatedly by a commercial CT imager at five different mAs levels from 100 down to 17 (the lowest value provided by the scanner). The preprocessed sinogram datasets were extracted from the CT scanner to a laboratory computer for noise analysis. The repeated measurements at each mAs level were used to test the normality of the repeatedly measured samples for each data channel using the Shapiro-Wilk statistical test merit. We further studied the probability distribution of the repeated measures. Most importantly, we validated a theoretical relationship between the sample mean and variance at each channel. It is our intention that the statistical test and particularly the relationship between the first and second statistical moments will improve low-dose CT image reconstruction for screening applications.

  13. Coherent scattering X-ray imaging at the Brazilian National Synchrotron Laboratory: Preliminary breast images

    Energy Technology Data Exchange (ETDEWEB)

    Castro, C.R.F. [Nuclear Instrumentation Laboratory-COPPE/UFRJ, P.O. Box 68509, Rio de Janeiro 21945-970 (Brazil); Barroso, R.C. [Physics Institute-University of Rio de Janeiro State, Rio de Janeiro 20559-900 (Brazil)]. E-mail: cely@uerij.br; Oliveira, L.F. de [Physics Institute-University of Rio de Janeiro State, Rio de Janeiro 20559-900 (Brazil); Lopes, R.T. [Nuclear Instrumentation Laboratory-COPPE/UFRJ, P.O. Box 68509, Rio de Janeiro 21945-970 (Brazil)

    2005-08-11

    The angular distribution of coherent scatter (low-momentum transfer) carries information about atomic structures, resulting in a pattern, which can be used to reconstruct a series of images. Coherent-scatter computed tomography is a novel imaging method developed to produce cross-sectional images based on the X-ray diffraction properties of an object. A different approach to coherent X-ray imaging is possible by fixing the detector at a given scatter angle {theta}, which produces an interference peak and then, carried out a tomography in the standard way. The cross-sectional images obtained allow determining the spatial dependence of coherent scatter cross-section of selected volume elements of inhomogeneous, extend objects for a single predetermined value of {theta} of interest, leading to a simplification of the data processing and the complexity of the apparatus. This work presents preliminary coherent scattering images carried out at the X-ray Diffraction beamline of the National Synchrotron Light Laboratory in Campinas, Brazil. The specimens were excised human breast tissues fixed in formaline. No frozen procedure was used in order to minimize preferred orientation during sample preparation. About 1mm thick slices cut from each of the fresh samples were mounted in frames without windows and placed on a translator to allow acquisition of scattering spectra. Cylinders containing healthy and cancerous (infiltrating ductal carcinoma) breast tissues were imagined at the characteristic angle for adipose tissue. Transmission and coherent scatter images are compared.

  14. Characterization of breast tissue using energy-dispersive X-ray diffraction computed tomography

    International Nuclear Information System (INIS)

    Pani, S.; Cook, E.J.; Horrocks, J.A.; Jones, J.L.; Speller, R.D.

    2010-01-01

    A method for sample characterization using energy-dispersive X-ray diffraction computed tomography (EDXRDCT) is presented. The procedures for extracting diffraction patterns from the data and the corrections applied are discussed. The procedures were applied to the characterization of breast tissue samples, 6 mm in diameter. Comparison with histological sections of the samples confirmed the possibility of grouping the patterns into five families, corresponding to adipose tissue, fibrosis, poorly differentiated cancer, well differentiated cancer and benign tumour.

  15. X-CT imaging method for large objects using double offset scan mode

    International Nuclear Information System (INIS)

    Fu Jian; Lu Hongnian; Li Bing; Zhang Lei; Sun Jingjing

    2007-01-01

    In X-ray computed tomography (X-CT) inspection, rotate-only scanner is commonly used because this configuration offers the highest imaging speed and best utilization of X-ray dose. But it requires that the imaging region of the scanned object must fit within the X-ray beam. Another configuration, transverse-rotate scanner, has a bigger field of view, but it is much more time consuming. In this paper, a two-dimensional X-CT imaging method for large objects is proposed to overcome the existing disadvantages. The scan principle of this method has been described and the reconstruction algorithm has been deduced. The results of the computer simulation and the experiments show the validity of the new method. Analysis shows that the scan field of view of this method is 1.8 times larger than that of rotate-only X-CT. The scan speed of this method is also much quicker than transverse-rotate X-CT

  16. Imaging of metastatic lymph nodes by X-ray phase-contrast micro-tomography.

    Directory of Open Access Journals (Sweden)

    Torben Haugaard Jensen

    Full Text Available Invasive cancer causes a change in density in the affected tissue, which can be visualized by x-ray phase-contrast tomography. However, the diagnostic value of this method has so far not been investigated in detail. Therefore, the purpose of this study was, in a blinded manner, to investigate whether malignancy could be revealed by non-invasive x-ray phase-contrast tomography in lymph nodes from breast cancer patients. Seventeen formalin-fixed paraffin-embedded lymph nodes from 10 female patients (age range 37-83 years diagnosed with invasive ductal carcinomas were analyzed by X-ray phase-contrast tomography. Ten lymph nodes had metastatic deposits and 7 were benign. The phase-contrast images were analyzed according to standards for conventional CT images looking for characteristics usually only visible by pathological examinations. Histopathology was used as reference. The result of this study was that the diagnostic sensitivity of the image analysis for detecting malignancy was 100% and the specificity was 87%. The positive predictive value was 91% for detecting malignancy and the negative predictive value was 100%. We conclude that x-ray phase-contrast imaging can accurately detect density variations to obtain information regarding lymph node involvement previously inaccessible with standard absorption x-ray imaging.

  17. WE-FG-207A-05: Dedicated Breast CT as a Diagnostic Imaging Tool: Physics and Clinical Feasibility

    International Nuclear Information System (INIS)

    Karellas, A.

    2016-01-01

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  18. WE-FG-207A-05: Dedicated Breast CT as a Diagnostic Imaging Tool: Physics and Clinical Feasibility

    Energy Technology Data Exchange (ETDEWEB)

    Karellas, A. [University of Massachusetts Medical School (United States)

    2016-06-15

    investigating dedicated breast CT. The development of large-area flat-panel detectors with field-of-view sufficient to image the entire breast in each projection enabled development of flat-panel cone-beam breast CT. More recently, the availability of complimentary metal-oxide semiconductor (CMOS) detectors with lower system noise and finer pixel pitch, combined with the development of x-ray tubes with focal spot dimensions similar to mammography systems, has shown improved spatial resolution and could improve visualization of microcalcifications. These technological developments promise clinical translation of low-dose cone-beam breast CT. Dedicated photon-counting breast CT (pcBCT) systems represent a novel detector design, which provide high spatial resolution (∼ 100µm) and low mean glandular dose (MGD). The CdTe-based direct conversion detector technology was previously evaluated and confirmed by simulations and basic experiments on laboratory setups [Kalender et al., Eur Radiol 22: 1–8, 2012]. Measurements of dose, technical image quality parameters, and surgical specimens on a pcBCT scanner have been completed. Comparative evaluation of surgical specimens showed that pcBCT outperformed mammography and digital breast tomosynthesis with respect to 3D spatial resolution, detectability of calcifications, and soft tissue delineation. Major barriers to widespread clinical use of BCT relate to radiation dose, imaging of microcalcifications, and adequate coverage of breast tissue near the chest wall. Adequate chest wall coverage is also technically challenging but recent progress in x-ray tube, detector and table design now enables full breast coverage in the majority of patients. At this time, BCT has been deemed to be suitable for diagnostic imaging but not yet for screening. The mean glandular dose (MGD) from BCT has been reported to be between 5.7 to 27.8 mGy, and this range is comparable to, and within the range of, the MGD of 2.6 to 31.6 mGy in diagnostic mammography

  19. Ceramic and polymeric devices for breast brachytherapy - Mammographic and CT response

    International Nuclear Information System (INIS)

    Nogueira, Luciana B.; Campos, Tarcisio P.R.

    2009-01-01

    The present research investigates the radiological visibility of ceramic and polymeric devices implanted in breast phantom (in vitro) for future applications in brachytherapy treatments. The main research goal is to investigate the viability of monitoring ceramic and polymeric devices, in vitro based on simple methods of radiological diagnostic, maintaining the easiest access to the population, represented by the conventional X-ray and mammography. The methodology involves the processing of ceramic devices constituted by bioglasses of Sm, SmBa, Ho, HoBa and the production of polymeric devices, such as polymeric membranes incorporating Ho e HoBa. Contrast agent of Barium was introduced in the syntheses of those devices to improve the radiological visibility in breast equivalent-tissue (TE) phantom. The breast phantom is constituted of glandular, adipose and skin TE, reproducing a 5 cm compressed real breast. In the compressed breast phantom, all types of ceramic and polymeric devices were implanted side by side. Radiological images were generated through X-ray equipment, mammography and computerized tomography (TC), for the samples implanted in the compressed breast phantom. The results show that SmBa and HoBa seeds on breast phantom presented suitable radiological visibility, on all the radiological diagnostic methods. However, the X-rays radiological visibility of Sm seeds without contrast was discreet. On mammography and TC images, it was not possible to identify those seeds, because the same ones were degraded after two months immersed in the glandular TE, after placed on the phantom. The Ho seeds were identified on all radiological diagnostic images, although non contrast agent in its constitution was added. However, the holmium polymeric membranes in direct contact with TE did not show Xray radiological visibility. However, the polymeric membranes of HoBa in the same conditions presented efficient X-rays radiological response. For mammography and TC methods

  20. 50 μm pixel pitch wafer-scale CMOS active pixel sensor x-ray detector for digital breast tomosynthesis

    International Nuclear Information System (INIS)

    Zhao, C; Kanicki, J; Konstantinidis, A C; Zheng, Y; Speller, R D; Anaxagoras, T

    2015-01-01

    Wafer-scale CMOS active pixel sensors (APSs) have been developed recently for x-ray imaging applications. The small pixel pitch and low noise are very promising properties for medical imaging applications such as digital breast tomosynthesis (DBT). In this work, we evaluated experimentally and through modeling the imaging properties of a 50 μm pixel pitch CMOS APS x-ray detector named DynAMITe (Dynamic Range Adjustable for Medical Imaging Technology). A modified cascaded system model was developed for CMOS APS x-ray detectors by taking into account the device nonlinear signal and noise properties. The imaging properties such as modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were extracted from both measurements and the nonlinear cascaded system analysis. The results show that the DynAMITe x-ray detector achieves a high spatial resolution of 10 mm −1 and a DQE of around 0.5 at spatial frequencies  <1 mm −1 . In addition, the modeling results were used to calculate the image signal-to-noise ratio (SNR i ) of microcalcifications at various mean glandular dose (MGD). For an average breast (5 cm thickness, 50% glandular fraction), 165 μm microcalcifications can be distinguished at a MGD of 27% lower than the clinical value (∼1.3 mGy). To detect 100 μm microcalcifications, further optimizations of the CMOS APS x-ray detector, image aquisition geometry and image reconstruction techniques should be considered. (paper)

  1. Observation of human tissue with phase-contrast x-ray computed tomography

    Science.gov (United States)

    Momose, Atsushi; Takeda, Tohoru; Itai, Yuji; Tu, Jinhong; Hirano, Keiichi

    1999-05-01

    Human tissues obtained from cancerous kidneys fixed in formalin were observed with phase-contrast X-ray computed tomography (CT) using 17.7-keV synchrotron X-rays. By measuring the distributions of the X-ray phase shift caused by samples using an X-ray interferometer, sectional images that map the distribution of the refractive index were reconstructed. Because of the high sensitivity of phase- contrast X-ray CT, a cancerous lesion was differentiated from normal tissue and a variety of other structures were revealed without the need for staining.

  2. 18F-FDG PET/CT for initial staging in breast cancer patients. Is there a relevant impact on treatment planning compared to conventional staging modalities?

    International Nuclear Information System (INIS)

    Krammer, J.; Schnitzer, A.; Kaiser, C.G.; Buesing, K.A.; Schoenberg, S.O.; Wasser, K.; Sperk, E.; Brade, J.; Wasgindt, S.; Suetterlin, M.; Sutton, E.J.

    2015-01-01

    To evaluate the impact of whole-body 18 F-FDG PET/CT on initial staging of breast cancer in comparison to conventional staging modalities. This study included 102 breast cancer patients, 101 patients were eligible for evaluation. Preoperative whole-body staging with PET/CT was performed in patients with clinical stage ≥ T2 tumours or positive local lymph nodes (n = 91). Postoperative PET/CT was performed in patients without these criteria but positive sentinel lymph node biopsy (n = 10). All patients underwent PET/CT and a conventional staging algorithm, which included bone scan, chest X-ray and abdominal ultrasound. PET/CT findings were compared to conventional staging and the impact on therapeutic management was evaluated. PET/CT led to an upgrade of the N or M stage in overall 19 patients (19 %) and newly identified manifestation of breast cancer in two patients (2 %). PET/CT findings caused a change in treatment of 11 patients (11 %). This is within the range of recent studies, all applying conventional inclusion criteria based on the initial T and N status. PET/CT has a relevant impact on initial staging and treatment of breast cancer when compared to conventional modalities. Further studies should assess inclusion criteria beyond the conventional T and N status, e.g. tumour grading and receptor status. (orig.)

  3. Development of a high-energy x-ray CT and its application to iron and steel analysis

    International Nuclear Information System (INIS)

    Taguchi, Isamu

    1987-01-01

    X-ray computed tomographic scanners are extensively used in medicine but have rarely been applied to non-medical purposes. Steel specimens pose particularly difficult problems - very poor transmission of X-rays and the need for high resolving capability. There have thus been no effective tomographic methods for examining steel specimens. Due to the growing need for non-destructive, non-contact methods for observing and analyzing the internal conditions of steel and raw materials for steel, however, we have developed a new high-energy computed tomographic scanner for steel (CTS). Its major specifications and functions are as follows. Type : 2nd-generation CT, 8-channel, Scanning method : 6deg revolution, 30-time traversing, Slice width : 0.3 mm, Resolving capability : 0.1 x 0.1 mm X-ray source : 420 kV, 3 mA, X-ray detector : BGO scintillator, Standard sample size : 50 mm dia., 50 mm high, Data collection time : 9.5 or 5 min. The CTS was successfully applied to the observation and the analysis of porosities of stainless steel (SUS 304) bloom, pores of iron ore sinters and metallic phases of the meteirites found in Antarctic Continent. (author)

  4. Computed tomography for light materials using a monochromatic X-ray beam produced by parametric X-ray radiation

    Energy Technology Data Exchange (ETDEWEB)

    Hayakawa, Y., E-mail: yahayak@lebra.nihon-u.ac.jp [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Hayakawa, K.; Inagaki, M. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Kaneda, T. [Nihon University School of Dentistry at Matsudo, Sakaecho-Nishi 2-870-1, Matsudo 271-8587 (Japan); Nakao, K.; Nogami, K. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Sakae, T. [Nihon University School of Dentistry at Matsudo, Sakaecho-Nishi 2-870-1, Matsudo 271-8587 (Japan); Sakai, T.; Sato, I. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan); Takahashi, Y. [Institute of Materials Structure Science, High Energy Accelerator Research Organization (KEK), Oho, Tsukuba 305-8501 (Japan); Tanaka, T. [Laboratory for Electron Beam Research and Application, Nihon University, Narashinodai 7-24-1, Funabashi 274-8501 (Japan)

    2013-08-15

    Computed tomography (CT) for light materials such as soft biological tissues was performed using a monochromatic X-ray beam provided by a parametric X-ray radiation (PXR) source at the Laboratory for Electron Beam Research and Application (LEBRA) of Nihon University. Using a high-efficiency flat panel detector (FPD), each projection image for CT was taken with exposure times of 5 or 10 s, and 60–360 projection images in each run were obtained with total measurement time of 5 min to 1 h. CT images were obtained from the projection images using the conventional calculation method. The typical tomograms obtained had sharp outlines, which are likely attributable to the propagation-based phase contrast.

  5. Application- and patient size-dependent optimization of x-ray spectra for CT

    International Nuclear Information System (INIS)

    Kalender, Willi A.; Deak, Paul; Kellermeier, Markus; Straten, Marcel van; Vollmar, Sabrina V.

    2009-01-01

    Although x-ray computed tomography (CT) has been in clinical use for over 3 decades, spectral optimization has not been a topic of great concern; high voltages around 120 kV have been in use since the beginning of CT. It is the purpose of this study to analyze, in a rigorous manner, the energies at which the patient dose necessary to provide a given contrast-to-noise ratio (CNR) for various diagnostic tasks can be minimized. The authors used cylindrical water phantoms and quasianthropomorphic phantoms of the thorax and the abdomen with inserts of 13 mm diameter mimicking soft tissue, bone, and iodine for simulations and measurements. To provide clearly defined contrasts, these inserts were made of solid water with a 1% difference in density (DD) to represent an energy-independent soft-tissue contrast of 10 Hounsfield units (HU), calcium hydroxyapatite (Ca) representing bone, and iodine (I) representing the typical contrast medium. To evaluate CT of the thorax, an adult thorax phantom (300x200 mm 2 ) plus extension rings up to a size of 460x300 mm 2 to mimic different patient cross sections were used. For CT of the abdomen, we used a phantom of 360x200 mm 2 and an extension ring of 460x300 mm 2 . The CT scanner that the authors used was a SOMATOM Definition (Siemens Healthcare, Forchheim, Germany) at 80, 100, 120, and 140 kV. Further voltage settings of 60, 75, 90, and 105 kV were available in an experimental mode. The authors determined contrast for the density difference, calcium, and iodine, and noise and 3D dose distributions for the available voltages by measurements. Additional voltage values and monoenergetic sources were evaluated by simulations. The dose-weighted contrast-to-noise ratio (CNRD) was used as the parameter for optimization. Simulations and measurements were in good agreement with respect to absolute values and trends regarding the dependence on energy for the parameters investigated. For soft-tissue imaging, the standard settings of 120-140 k

  6. Patient dose simulation in X-ray CT using a radiation treatment-planning system

    International Nuclear Information System (INIS)

    Nakae, Yasuo; Oda, Masahiko; Minamoto, Takahiro

    2003-01-01

    Medical irradiation dosage has been increasing with the development of new radiological equipment and new techniques like interventional radiology. It is fair to say that patient dose has been increased as a result of the development of multi-slice CT. A number of studies on the irradiation dose of CT have been reported, and the computed tomography dose index (CTDI) is now used as a general means of determining CT dose. However, patient dose distribution in the body varies with the patient's constitution, bowel gas in the body, and conditions of exposure. In this study, patient dose was analyzed from the viewpoint of dose distribution, using a radiation treatment-planning computer. Percent depth dose (PDD) and the off-center ratio (OCR) of the CT beam are needed to calculate dose distribution by the planning computer. Therefore, X-ray CT data were measured with various apparatuses, and beam data were sent to the planning computer. Measurement and simulation doses in the elliptical phantom (Mix-Dp: water equivalent material) were collated, and the CT irradiation dose was determined for patient dose simulation. The rotational radiation treatment technique was used to obtain the patient dose distribution of CT, and patient dose was evaluated through simulation of the dose distribution. CT images of the thorax were sent to the planning computer and simulated. The result was that the patient dose distribution of the thorax was obtained for CT examination. (author)

  7. WE-AB-BRA-11: Improved Imaging of Permanent Prostate Brachytherapy Seed Implants by Combining an Endorectal X-Ray Sensor with a CT Scanner

    International Nuclear Information System (INIS)

    Steiner, J; Matthews, K; Jia, G

    2016-01-01

    Purpose: To test feasibility of the use of a digital endorectal x-ray sensor for improved image resolution of permanent brachytherapy seed implants compared to conventional CT. Methods: Two phantoms simulating the male pelvic region were used to test the capabilities of a digital endorectal x-ray sensor for imaging permanent brachytherapy seed implants. Phantom 1 was constructed from acrylic plastic with cavities milled in the locations of the prostate and the rectum. The prostate cavity was filled a Styrofoam plug implanted with 10 training seeds. Phantom 2 was constructed from tissue-equivalent gelatins and contained a prostate phantom implanted with 18 strands of training seeds. For both phantoms, an intraoral digital dental x-ray sensor was placed in the rectum within 2 cm of the seed implants. Scout scans were taken of the phantoms over a limited arc angle using a CT scanner (80 kV, 120–200 mA). The dental sensor was removed from the phantoms and normal helical CT and scout (0 degree) scans using typical parameters for pelvic CT (120 kV, auto-mA) were collected. A shift-and add tomosynthesis algorithm was developed to localize seed plane location normal to detector face. Results: The endorectal sensor produced images with improved resolution compared to CT scans. Seed clusters and individual seed geometry were more discernable using the endorectal sensor. Seed 3D locations, including seeds that were not located in every projection image, were discernable using the shift and add algorithm. Conclusion: This work shows that digital endorectal x-ray sensors are a feasible method for improving imaging of permanent brachytherapy seed implants. Future work will consist of optimizing the tomosynthesis technique to produce higher resolution, lower dose images of 1) permanent brachytherapy seed implants for post-implant dosimetry and 2) fine anatomic details for imaging and managing prostatic disease compared to CT images. Funding: LSU Faculty Start-up Funding

  8. WE-AB-BRA-11: Improved Imaging of Permanent Prostate Brachytherapy Seed Implants by Combining an Endorectal X-Ray Sensor with a CT Scanner

    Energy Technology Data Exchange (ETDEWEB)

    Steiner, J; Matthews, K; Jia, G [Louisiana State University, Baton Rouge, LA (United States)

    2016-06-15

    Purpose: To test feasibility of the use of a digital endorectal x-ray sensor for improved image resolution of permanent brachytherapy seed implants compared to conventional CT. Methods: Two phantoms simulating the male pelvic region were used to test the capabilities of a digital endorectal x-ray sensor for imaging permanent brachytherapy seed implants. Phantom 1 was constructed from acrylic plastic with cavities milled in the locations of the prostate and the rectum. The prostate cavity was filled a Styrofoam plug implanted with 10 training seeds. Phantom 2 was constructed from tissue-equivalent gelatins and contained a prostate phantom implanted with 18 strands of training seeds. For both phantoms, an intraoral digital dental x-ray sensor was placed in the rectum within 2 cm of the seed implants. Scout scans were taken of the phantoms over a limited arc angle using a CT scanner (80 kV, 120–200 mA). The dental sensor was removed from the phantoms and normal helical CT and scout (0 degree) scans using typical parameters for pelvic CT (120 kV, auto-mA) were collected. A shift-and add tomosynthesis algorithm was developed to localize seed plane location normal to detector face. Results: The endorectal sensor produced images with improved resolution compared to CT scans. Seed clusters and individual seed geometry were more discernable using the endorectal sensor. Seed 3D locations, including seeds that were not located in every projection image, were discernable using the shift and add algorithm. Conclusion: This work shows that digital endorectal x-ray sensors are a feasible method for improving imaging of permanent brachytherapy seed implants. Future work will consist of optimizing the tomosynthesis technique to produce higher resolution, lower dose images of 1) permanent brachytherapy seed implants for post-implant dosimetry and 2) fine anatomic details for imaging and managing prostatic disease compared to CT images. Funding: LSU Faculty Start-up Funding

  9. Fundamentals for cost calculations of X-ray equipment

    International Nuclear Information System (INIS)

    Bossard, F.

    1985-01-01

    Economic implications of running an X-ray departement in Switzerland will be illustrated by comparing operating costs of private radiological institutes with and without CT to the operating costs of large radiological departments in hospitals with and without CT and to the operating costs of simple X-ray equipment in general practicioners' offices. - These costs calculations form the basis for cost-benefit analyses. (orig.) [de

  10. X-ray induced formation of γ-H2AX foci after full-field digital mammography and digital breast-tomosynthesis.

    Directory of Open Access Journals (Sweden)

    Siegfried A Schwab

    Full Text Available PURPOSE: To determine in-vivo formation of x-ray induced γ-H2AX foci in systemic blood lymphocytes of patients undergoing full-field digital mammography (FFDM and to estimate foci after FFDM and digital breast-tomosynthesis (DBT using a biological phantom model. MATERIALS AND METHODS: The study complies with the Declaration of Helsinki and was performed following approval by the ethic committee of the University of Erlangen-Nuremberg. Written informed consent was obtained from every patient. For in-vivo tests, systemic blood lymphocytes were obtained from 20 patients before and after FFDM. In order to compare in-vivo post-exposure with pre-exposure foci levels, the Wilcoxon matched pairs test was used. For in-vitro experiments, isolated blood lymphocytes from healthy volunteers were irradiated at skin and glandular level of a porcine breast using FFDM and DBT. Cells were stained against the phosphorylated histone variant γ-H2AX, and foci representing distinct DNA damages were quantified. RESULTS: Median in-vivo foci level/cell was 0.086 (range 0.067-0.116 before and 0.094 (0.076-0.126 after FFDM (p = 0.0004. In the in-vitro model, the median x-ray induced foci level/cell after FFDM was 0.120 (range 0.086-0.140 at skin level and 0.035 (range 0.030-0.050 at glandular level. After DBT, the median x-ray induced foci level/cell was 0.061 (range 0.040-0.081 at skin level and 0.015 (range 0.006-0.020 at glandular level. CONCLUSION: In patients, mammography induces a slight but significant increase of γ-H2AX foci in systemic blood lymphocytes. The introduced biological phantom model is suitable for the estimation of x-ray induced DNA damages in breast tissue in different breast imaging techniques.

  11. Elastographic versus x-ray CT imaging of radio frequency ablation coagulations: An in vitro study

    International Nuclear Information System (INIS)

    Liu Wu; Techavipoo, Udomchai; Varghese, Tomy; Zagzebski, James A.; Chen, Quan; Lee, Fred T. Jr.

    2004-01-01

    Techniques to image elasticity parameters (i.e., elastography) have recently become of great interest to researchers. In this paper we use conventional ultrasound elastography and x-ray CT to image radio frequency (RF) ablation sites of excised canine liver enclosed in gelatin. Thermal coagulations of different sizes were produced by applying the RF procedure for various times and end point temperatures. Dimensions, areas and volumes computed from CT and elastography were compared with those on whole mount pathology specimens. Ultrasound elastography exhibited high contrast for the thermal coagulations and performed better than CT. The correlation between pathology and elastography for this sample set of 40 thermal coagulations (r=0.94 for volume estimation, r=0.87 for area estimation) is better than the correlation between pathology and CT (r=0.89 for volume estimation, r=0.82 for area estimation)

  12. Variance analysis of x-ray CT sinograms in the presence of electronic noise background.

    Science.gov (United States)

    Ma, Jianhua; Liang, Zhengrong; Fan, Yi; Liu, Yan; Huang, Jing; Chen, Wufan; Lu, Hongbing

    2012-07-01

    Low-dose x-ray computed tomography (CT) is clinically desired. Accurate noise modeling is a fundamental issue for low-dose CT image reconstruction via statistics-based sinogram restoration or statistical iterative image reconstruction. In this paper, the authors analyzed the statistical moments of low-dose CT data in the presence of electronic noise background. The authors first studied the statistical moment properties of detected signals in CT transmission domain, where the noise of detected signals is considered as quanta fluctuation upon electronic noise background. Then the authors derived, via the Taylor expansion, a new formula for the mean-variance relationship of the detected signals in CT sinogram domain, wherein the image formation becomes a linear operation between the sinogram data and the unknown image, rather than a nonlinear operation in the CT transmission domain. To get insight into the derived new formula by experiments, an anthropomorphic torso phantom was scanned repeatedly by a commercial CT scanner at five different mAs levels from 100 down to 17. The results demonstrated that the electronic noise background is significant when low-mAs (or low-dose) scan is performed. The influence of the electronic noise background should be considered in low-dose CT imaging.

  13. X-ray production experiments on the RACE Compact Torus Accelerator

    International Nuclear Information System (INIS)

    Hammer, J.H.; Eddleman, J.L.; Hartman, C.W.; McLean, H.S.; Molvik, A.W.; Gee, M.

    1989-12-01

    The Purpose of the Compact Torus Accelerator (CTA) program at LLNL is to prove the principle of a unique accelerator concept based on magnetically confined compact torus (CT) plasma rings and to study applications. Successful achievement of these goals could lead to a high power-density driver for many applications including an intense x-ray source for nuclear weapons effects simulation and an inertial fusion driver. Fusion applications and a description of the CTA concept are included in a companion paper at this conference. This paper will describe the initial experiments on soft x-ray production conducted on the plasma Ring ACcelerator Experiment (RACE) and compare the results to modeling studies. The experiments on CT stagnation and soft x-ray production were conducted with unfocused rings as a first of CT dynamics and the physics of x-ray production. The x-ray fluences observed are consistent with expectations based on calculations employing a radiation-hydrodynamics code. We conclude with a diffusion of future x-ray production studies that can be conducted on RACE and a possible multi-megajoule upgrade

  14. Quantitative measurement of lung density with x-ray CT and positron CT, (2). Diseased subjects

    Energy Technology Data Exchange (ETDEWEB)

    Ito, Kengo; Ito, Masatoshi; Kubota, Kazuo

    1985-05-01

    Lung density was quantitatively measured on six diseased patients with X-ray CT (XCT) and Positron CT(PCT). The findings are as follows: In the silicosis, extravascular lung density was found to be remarkably increased compared to normals (0.29gcm/sup 3/), but blood volume was in normal range. In the post-irradiated lung cancers, extravascular lung density increased in the irradiated sites compared to the non-irradiated opposite sites, and blood volume varied in each case. In a patient with chronic heart failure, blood volume decreased (0.11mlcm/sup 3/) with increased extravascular lung density (0.23gcm/sup 3/). In the chronic obstructive pulmonary disease, both extravascular lung density and blood volume decreased (0.11gcm/sup 3/ and 0.10mlcm/sup 3/ respectively). Lung density measured with XCT was constantly lower than that with PCT in all cases. But changes in the values of lung density measured, correlated well with each other. In conclusion, the method presented here may clarify the etiology of the diffuse pulmonary diseases, and be used to differentiate and grade the diseases.

  15. TH-AB-209-11: Breast Microcalcification Classification Using Spectral X-Ray Coherent Scatter Computed Tomography

    Energy Technology Data Exchange (ETDEWEB)

    Ghammraoui, B; M Popescu, L; Badano, A [Food & Drug Administration, Silver Spring, MD (United States)

    2016-06-15

    Purpose: To investigate the ability of Coherent Scatter Computed Tomography (CSCT) to distinguish non-invasively between type I calcifications, consisting of calcium oxalate dihydrate (CO) compounds which are more often associated with benign lesions, and type II calcifications containing hydroxyapatite (HA) which are predominantly associated with malignant tumors. Methods: The coherent scatter cross sections of HA and CO were measured using an energy dispersive x-ray diffractometer. The measured cross sections were introduced into MC-GPU Monte Carlo simulation code for studying the applicability of CSCT to discriminate between the two types of microcalcifications within the whole breast. Simulations were performed on a virtual phantom with inserted HA and CO spots of different sizes and placed in regions of interest having different background compositions. We considered a polychromatic x-ray source and an energy resolving photon counting detector. We applied an algorithm that estimates scatter components in projection space in order to obtain material-specific images of the breast. As material components adipose, glandular, HA and CO were used. The relative contrast of HA and CO components were used for type I and type II microcalcification discrimination. Results: The reconstructed CSCT images showed material-specific component-contrast values, with the highest CO or HA component contrast corresponding generally to the actual CO or HA feature, respectively. The discrimination performance varies with the x-ray intensity, calcification size, and background composition. The results were summarized using receiver operating characteristic (ROC) analysis with the area under the curve (AUC) taken as an overall indicator of discrimination performance and showing high AUC values up to unity. Conclusion: The simulation results obtained for a uniform breast imaging phantom indicate that CSCT has potential to be used as a non-invasive method for discrimination between type

  16. X-ray diagnosis in temporal bone anomalies

    International Nuclear Information System (INIS)

    Schratter, M.; Canigiani, G.; Swoboda, H.; Brunner, E.

    1988-01-01

    The X-ray findings in temporal bone anomalies are reviewed. Radiological procedure and examination technique are presented, as are symptoms of important anomalies. The methods available are plain film X-ray of temporal bone, multi-directional tomography, and high-resolution CT. Although some of the abnormalities are visible even in plain films, consistent use of conventional tomography or CT is necessary for correct diagnosis. This procedure is indicated not only when an abnormality is clinically obvious, but also in all cases of unexplained hearing loss without evidence of acquired disease. The advantage of CT over conventional tomography is that soft tissue anomalies, such as primary cholesteatoma or tumor simulating vascular abnormalities, can be demonstrated. In these cases CT is obligatory. (orig.) [de

  17. Evaluation of patient dose in imaging using a cone-beam CT dosimetry by X-ray films for radiotherapeutic dose

    International Nuclear Information System (INIS)

    Yoshida, Yuri; Morita, Yasuhiko; Honda, Eiichi; Tomotake, Yoritoki; Ichikawa, Tetsuo

    2008-01-01

    A limited cone-beam X-ray CT (3DX multi-image micro CT; 3DX-FPD) is widely used in dentistry because it provides a lower cost, smaller size, and higher spatial resolution than a CT for medicine. Our recent research suggested that the patient dose of 3DX-FPD was less than 7/10 of that of CT, and it was several to 10 times more than that of dental or panoramic radiography. The purpose of this study was to evaluate the spatial dose distribution from 3DX-FPD and to estimate the influence of dose by positioning of the region of interest. Dosimetry of the organs and the tissues was performed using an anthropomorphic Alderson Rando phantom and X-ray films for measurement of radiotherapeutic dose. Measurements of dose distribution were performed using a cylinder-type tank of water made of acrylic resin imitating the head and X-ray films. The results are summarized as follows: The dose was higher as the ratio of the air region included in the region of interest increased. The dose distribution was not homogeneous and the dose was highest in the skin region. The dose was higher for several seconds after the beginning of exposure. It was concluded that patient positioning, as well as exposure conditions including the size of the exposure field and tube current, could greatly influence the patient dose in 3DX-FPD. In addition, it is necessary to consider the influence of image quality for the treatment of dental implants. (author)

  18. Laboratory based study of dynamical processes by 4D X-ray CT with sub-second temporal resolution

    Czech Academy of Sciences Publication Activity Database

    Vavřík, Daniel; Jakůbek, Jan; Kumpová, Ivana; Pichotka, M.

    2017-01-01

    Roč. 12, February (2017), č. článku C02010. ISSN 1748-0221. [International Workshop on Radiation Imaging Detectors /18./. Barcelona, 03.07.2016-07.07.2016] R&D Projects: GA ČR(CZ) GA15-07210S; GA MŠk(CZ) LO1219 Keywords : computerized tomography (CT) * computed radiography (CR) * inspection with x-rays * pixelated detectors and associated VLSI electronics * X-ray radiography and digital radiography (DR) Subject RIV: JB - Sensors, Measurment, Regulation OBOR OECD: Electrical and electronic engineering Impact factor: 1.220, year: 2016 http://iopscience.iop.org/article/10.1088/1748-0221/12/02/C02010

  19. X-ray and gamma-ray transmission computed tomographic imaging of archaeological objects

    International Nuclear Information System (INIS)

    Jaafar Abdullah; Susan Maria Sipaun

    2004-01-01

    X-ray or gamma-ray transmission computed tomography (CT) is a powerful non-destructive evaluation (NDE) technique that produces two-dimensional cross-section images of an object without the need to physically section it. CT is also known by the acronym CAT, for computerised axial tomography or computed-aided tomography. The invention of CT techniques revolutionised the field of medical diagnostic imaging because it provided more detailed and useful information than any previous non-invasive imaging techniques. The method is increasingly being used in industry, aerospace, geosciences and archaeology. This paper presents a brief overview of X-ray or gamma-ray transmission tomography. It is not intended to be a technical treatise but is hoped that it would raise awareness and promote opportunities for further collaboration amongst the nuclear research community, including archaeologists and those in the conservation profession. The theoretical aspects of CT scanner, the system configurations and the adopted algorithm for image reconstruction are discussed. In addition, a few examples of CT images for archaeological objects are presented. The examples were purposely chosen to illustrate clearly and precisely the fundamental concepts of this sophisticated field. (Author)

  20. Radiation doses in volume-of-interest breast computed tomography—A Monte Carlo simulation study

    Energy Technology Data Exchange (ETDEWEB)

    Lai, Chao-Jen, E-mail: cjlai3711@gmail.com; Zhong, Yuncheng; Yi, Ying; Wang, Tianpeng; Shaw, Chris C. [Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030-4009 (United States)

    2015-06-15

    Purpose: Cone beam breast computed tomography (breast CT) with true three-dimensional, nearly isotropic spatial resolution has been developed and investigated over the past decade to overcome the problem of lesions overlapping with breast anatomical structures on two-dimensional mammographic images. However, the ability of breast CT to detect small objects, such as tissue structure edges and small calcifications, is limited. To resolve this problem, the authors proposed and developed a volume-of-interest (VOI) breast CT technique to image a small VOI using a higher radiation dose to improve that region’s visibility. In this study, the authors performed Monte Carlo simulations to estimate average breast dose and average glandular dose (AGD) for the VOI breast CT technique. Methods: Electron–Gamma-Shower system code-based Monte Carlo codes were used to simulate breast CT. The Monte Carlo codes estimated were validated using physical measurements of air kerma ratios and point doses in phantoms with an ion chamber and optically stimulated luminescence dosimeters. The validated full cone x-ray source was then collimated to simulate half cone beam x-rays to image digital pendant-geometry, hemi-ellipsoidal, homogeneous breast phantoms and to estimate breast doses with full field scans. 13-cm in diameter, 10-cm long hemi-ellipsoidal homogeneous phantoms were used to simulate median breasts. Breast compositions of 25% and 50% volumetric glandular fractions (VGFs) were used to investigate the influence on breast dose. The simulated half cone beam x-rays were then collimated to a narrow x-ray beam with an area of 2.5 × 2.5 cm{sup 2} field of view at the isocenter plane and to perform VOI field scans. The Monte Carlo results for the full field scans and the VOI field scans were then used to estimate the AGD for the VOI breast CT technique. Results: The ratios of air kerma ratios and dose measurement results from the Monte Carlo simulation to those from the physical

  1. Human soft tissue analysis using x-ray or gamma-ray techniques

    International Nuclear Information System (INIS)

    Theodorakou, C; Farquharson, M J

    2008-01-01

    This topical review is intended to describe the x-ray techniques used for human soft tissue analysis. X-ray techniques have been applied to human soft tissue characterization and interesting results have been presented over the last few decades. The motivation behind such studies is to provide improved patient outcome by using the data obtained to better understand a disease process and improve diagnosis. An overview of theoretical background as well as a complete set of references is presented. For each study, a brief summary of the methodology and results is given. The x-ray techniques include x-ray diffraction, x-ray fluorescence, Compton scattering, Compton to coherent scattering ratio and attenuation measurements. The soft tissues that have been classified using x-rays or gamma rays include brain, breast, colon, fat, kidney, liver, lung, muscle, prostate, skin, thyroid and uterus. (topical review)

  2. Performance evaluation of 3-D enhancement filters for detection of lung cancer from 3-D chest X-ray CT images

    International Nuclear Information System (INIS)

    Shimizu, Akinobu; Hagai, Makoto; Toriwaki, Jun-ichiro; Hasegawa, Jun-ichi.

    1995-01-01

    This paper evaluates the performance of several three dimensional enhancement filters used in procedures for detecting lung cancer shadows from three dimensional (3D) chest X-ray CT images. Two dimensional enhancement filters such as Min-DD filter, Contrast filter and N-Quoit filter have been proposed for enhancing cancer shadows in conventional 2D X-ray images. In this paper, we extend each of these 2D filters to a 3D filter and evaluate its performance experimentally by using CT images with artificial and true lung cancer shadows. As a result, we find that these 3D filters are effective for determining the position of a lung cancer shadow in a 3D chest CT image, as compared with the simple procedure such as smoothing filter, and that the performance of these filters become lower in the hilar area due to the influence of the vessel shadows. (author)

  3. Gamma-ray detectors for breast imaging

    Science.gov (United States)

    Williams, Mark B.; Goode, Allen R.; Majewski, Stan; Steinbach, Daniela; Weisenberger, Andrew G.; Wojcik, Randolph F.; Farzanpay, Farzin

    1997-07-01

    Breast cancer is the most common cancer of American women and is the leading cause of cancer-related death among women aged 15 - 54; however recent years have shown that early detection using x-ray mammography can lead to a high probability of cure. However, because of mammography's low positive predictive value, surgical or core biopsy is typically required for diagnosis. In addition, the low radiographic contrast of many nonpalpable breast masses, particularly among women with radiographically dense breasts, results in an overall rate of 10% to 25% for missed tumors. Nuclear imaging of the breast using single gamma emitters (scintimammography) such as (superscript 99m)Tc, or positron emitters such as F-18- fluorodeoxyglucose (FDG) for positron emission tomography (PET), can provide information on functional or metabolic tumor activity that is complementary to the structural information of x-ray mammography, thereby potentially reducing the number of unnecessary biopsies and missed cancers. This paper summarizes recent data on the efficacy of scintimammography using conventional gamma cameras, and describes the development of dedicated detectors for gamma emission breast imaging. The detectors use new, high density crystal scintillators and large area position sensitive photomultiplier tubes (PSPMTs). Detector design, imaging requirements, and preliminary measured imaging performance are discussed.

  4. Explaining Air and Water Transport in Undisturbed Soils By X-Ray CT Derived Macroporosity and CT- Number-Derived Matrix Density

    DEFF Research Database (Denmark)

    Paradelo Pérez, Marcos; Katuwal, Sheela; Møldrup, Per

    The characterization of soil pore space geometry is important to predict the fluxes of air, water and solutes through soil and understand soil hydrogeochemical functions. X-ray computed tomography (CT) -derived parameters were evaluated as predictors of water, air and solute transport through soil....... Forty five soil columns (20-cm × 20-cm) were collected at an agricultural field in Estrup, Denmark. The soil columns were scanned in a medical CT-scanner. Subsequent to this, non-reactive tracer leaching experiments were performed in the laboratory together with measurements of air permeability (Ka...... is considered a robust indicator of preferential flow. Meanwhile, CT-derived limiting macro-porosity was the best predictor for Ka and log10Ksat. A best subsets regression analysis was performed combining macroporosity, limiting macroporosity and CTmatrix. The predictions of water and air flow improved using...

  5. Characteristic performance evaluation of a photon counting Si strip detector for low dose spectral breast CT imaging

    Science.gov (United States)

    Cho, Hyo-Min; Barber, William C.; Ding, Huanjun; Iwanczyk, Jan S.; Molloi, Sabee

    2014-01-01

    Purpose: The possible clinical applications which can be performed using a newly developed detector depend on the detector's characteristic performance in a number of metrics including the dynamic range, resolution, uniformity, and stability. The authors have evaluated a prototype energy resolved fast photon counting x-ray detector based on a silicon (Si) strip sensor used in an edge-on geometry with an application specific integrated circuit to record the number of x-rays and their energies at high flux and fast frame rates. The investigated detector was integrated with a dedicated breast spectral computed tomography (CT) system to make use of the detector's high spatial and energy resolution and low noise performance under conditions suitable for clinical breast imaging. The aim of this article is to investigate the intrinsic characteristics of the detector, in terms of maximum output count rate, spatial and energy resolution, and noise performance of the imaging system. Methods: The maximum output count rate was obtained with a 50 W x-ray tube with a maximum continuous output of 50 kVp at 1.0 mA. A109Cd source, with a characteristic x-ray peak at 22 keV from Ag, was used to measure the energy resolution of the detector. The axial plane modulation transfer function (MTF) was measured using a 67 μm diameter tungsten wire. The two-dimensional (2D) noise power spectrum (NPS) was measured using flat field images and noise equivalent quanta (NEQ) were calculated using the MTF and NPS results. The image quality parameters were studied as a function of various radiation doses and reconstruction filters. The one-dimensional (1D) NPS was used to investigate the effect of electronic noise elimination by varying the minimum energy threshold. Results: A maximum output count rate of 100 million counts per second per square millimeter (cps/mm2) has been obtained (1 million cps per 100 × 100 μm pixel). The electrical noise floor was less than 4 keV. The energy resolution

  6. Instruments for radiation measurement in life sciences (5). 'Development of imaging Technology in life sciences'. 5. X-ray CT for laboratory animals

    International Nuclear Information System (INIS)

    Tamegai, Toshiaki

    2007-01-01

    X-ray computed tomography, commercialized by EMI Co., UK, in 1973 and now used world-widely, is used not only for medical use but also for laboratory animals such as rats and mice to measure bone density and to obtain fine structures of bones. This paper introduces X-ray CT apparatus specifically designed for laboratory animals. Besides general explanations about the method, followed by emphasis on important performance of the measuring system, the paper explains technical aspects for obtaining the CT imaging scan procedure thus showing several photographs as example and introducing some clinical applications. (S. Ohno)

  7. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... bear denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams ... the web pages found at these links. About Us | Contact Us | FAQ | Privacy | Terms of Use | Links | ...

  8. Bone X-Ray (Radiography)

    Medline Plus

    Full Text Available ... fracture. guide orthopedic surgery, such as spine repair/fusion, joint replacement and fracture reductions. look for injury, ... CT Exams Arthritis X-ray, Interventional Radiology and Nuclear Medicine Radiation Safety How to Read Your Radiology ...

  9. Tc-99m Methoxy Isobutyl Isonitrile (MIBI) scintimammography for detection of loco-regional recurrence of breast cancer: A comparative study of scintimammography, X-ray mammography and breast ultrasound imaging

    International Nuclear Information System (INIS)

    Usmani, S.; Niaz, K.; Ahmed, N.; Niyaz, K.; Khan, H.A.; Javed, A.; Mohannadi, S.A.; Huda, F.A.

    2007-01-01

    Breast cancer is the most common cancer among women and the second leading cause of death in women after lung cancer. Local recurrence rate is about 1-2% per year. It is important to find a sensitive and accurate diagnostic method for early detection of recurrent tumor for planning subsequent management. The principal aim of the present study was to assess the utility of Tc-99m MIBI scintimammography in the detection of loco-regional recurrence of breast cancer in comparison with the conventionally used methods like x-ray mammography and ultrasonography. A total number of 42 patients (Age: mean= 45.55 yrs; median = 43.5; range, 22-77 yrs) suspected to have loco-regional recurrence of breast cancer on the basis of clinical examination and/or conventional imaging procedures were included in this study. All patients were subjected to planar scintimammography (SMM), SPECT imaging of the breast, x-ray mammography and Ultrasound scanning. The results were correlated with the final diagnosis arrived at by fine needle aspiration cytology (FNAC), core biopsy or definite wide local excision. Sensitivity (SEN), specificity (SPE), positive predictive value (PPV), negative predictive value (NPV) and accuracy (ACC) of scintimammography in the diagnosis of recurrent disease were found to be 82.75, 92.3, 96.0, 70.58 and 85.71% respectively; compared to 45.83, 66.6, 78.57, 31.57, 51.51 for x-ray mammography and 65.51, 69.23, 82.6, 47.36, 66.66 for Ultrasonography. Scintimammography results were correlated with x-ray mammography and ultrasonography by applying Pearson correlation. No significant correlation was found between SMM and XMM (r=0.27) or SMM and US (r=0.23). Based on the results of this study it is concluded that SMM has better diagnostic accuracy than mammography and ultrasonography in the detection of loco-regional recurrence of breast cancer. (author)

  10. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.

    Science.gov (United States)

    Kinahan, Paul E; Hasegawa, Bruce H; Beyer, Thomas

    2003-07-01

    A synergy of positron emission tomography (PET)/computed tomography (CT) scanners is the use of the CT data for x-ray-based attenuation correction of the PET emission data. Current methods of measuring transmission use positron sources, gamma-ray sources, or x-ray sources. Each of the types of transmission scans involves different trade-offs of noise versus bias, with positron transmission scans having the highest noise but lowest bias, whereas x-ray scans have negligible noise but the potential for increased quantitative errors. The use of x-ray-based attenuation correction, however, has other advantages, including a lack of bias introduced from post-injection transmission scanning, which is an important practical consideration for clinical scanners, as well as reduced scan times. The sensitivity of x-ray-based attenuation correction to artifacts and quantitative errors depends on the method of translating the CT image from the effective x-ray energy of approximately 70 keV to attenuation coefficients at the PET energy of 511 keV. These translation methods are usually based on segmentation and/or scaling techniques. Errors in the PET emission image arise from positional mismatches caused by patient motion or respiration differences between the PET and CT scans; incorrect calculation of attenuation coefficients for CT contrast agents or metallic implants; or keeping the patient's arms in the field of view, which leads to truncation and/or beam-hardening (or x-ray scatter) artifacts. Proper interpretation of PET emission images corrected for attenuation by using the CT image relies on an understanding of the potential artifacts. In cases where an artifact or bias is suspected, careful inspection of all three available images (CT and PET emission with and without attenuation correction) is recommended. Copyright 2003 Elsevier Inc. All rights reserved.

  11. X-ray diagnostics - benefits and risks; Roentgendiagnostik - Nutzen und Risiken

    Energy Technology Data Exchange (ETDEWEB)

    Bartholomaeus, Melanie (comp.)

    2016-10-15

    The brochure on benefits and risks of X-ray diagnostics discusses the following issues: X radiation - a pioneering discovery and medical sensation, fundamentals of X radiation, frequency of X-ray examinations in Germany in relation to CT imaging, radiation doses resulting from X-ray diagnostics, benefits of X-ray diagnostics - indication and examples, risks - measures for radiation exposure reductions, avoidance of unnecessary examinations.

  12. Systematic errors in digital volume correlation due to the self-heating effect of a laboratory x-ray CT scanner

    International Nuclear Information System (INIS)

    Wang, B; Pan, B; Tao, R; Lubineau, G

    2017-01-01

    The use of digital volume correlation (DVC) in combination with a laboratory x-ray computed tomography (CT) for full-field internal 3D deformation measurement of opaque materials has flourished in recent years. During x-ray tomographic imaging, the heat generated by the x-ray tube changes the imaging geometry of x-ray scanner, and further introduces noticeable errors in DVC measurements. In this work, to provide practical guidance high-accuracy DVC measurement, the errors in displacements and strains measured by DVC due to the self-heating for effect of a commercially available x-ray scanner were experimentally investigated. The errors were characterized by performing simple rescan tests with different scan durations. The results indicate that the maximum strain errors associated with the self-heating of the x-ray scanner exceed 400 µε . Possible approaches for minimizing or correcting these displacement and strain errors are discussed. Finally, a series of translation and uniaxial compression tests were performed, in which strain errors were detected and then removed using pre-established artificial dilatational strain-time curve. Experimental results demonstrate the efficacy and accuracy of the proposed strain error correction approach. (paper)

  13. Systematic errors in digital volume correlation due to the self-heating effect of a laboratory x-ray CT scanner

    KAUST Repository

    Wang, B

    2017-02-15

    The use of digital volume correlation (DVC) in combination with a laboratory x-ray computed tomography (CT) for full-field internal 3D deformation measurement of opaque materials has flourished in recent years. During x-ray tomographic imaging, the heat generated by the x-ray tube changes the imaging geometry of x-ray scanner, and further introduces noticeable errors in DVC measurements. In this work, to provide practical guidance high-accuracy DVC measurement, the errors in displacements and strains measured by DVC due to the self-heating for effect of a commercially available x-ray scanner were experimentally investigated. The errors were characterized by performing simple rescan tests with different scan durations. The results indicate that the maximum strain errors associated with the self-heating of the x-ray scanner exceed 400 µε. Possible approaches for minimizing or correcting these displacement and strain errors are discussed. Finally, a series of translation and uniaxial compression tests were performed, in which strain errors were detected and then removed using pre-established artificial dilatational strain-time curve. Experimental results demonstrate the efficacy and accuracy of the proposed strain error correction approach.

  14. Efficacy of Breast Shielding During CT of the Head

    International Nuclear Information System (INIS)

    Brnic, Z.; Vekic, B.; Hebrang, A.; Anic, P.

    2003-01-01

    The use of computerized tomography (CT) is rapidly increasing in last two decades, and this method has become the major non-natural source of radiation exposure to the population. CT examinations delivers to the patients more radiation than all other imaging techniques, and contribute disproportionately to the collective dose; in Britain it has been estimated that 4% of diagnostic radiology procedures are CT examinations, being responsible for approximately 40% of the total annual collective dose. Breast doses are high in CT examinations with breasts in scanning planes, being not insignificant also when breasts are exposed only to scatter radiation. Breast doses received through scatter radiation during head CT may account for up to one-fifth of an average mammographic dose per one view. Whilst the possibilities of reduction of radiation load to organs lying in CT scanning planes are limited, the tissues outside the primary beam should be protected against scatter whenever it does not sacrifice image quality. Lead shielding results in significant reduction of external scatter to radiosensitive superficial organs in many diagnostic procedures. The published studies of breast shielding against scatter radiation in diagnostic radiology are scanty, only the later one dealing particularly with breast shielding during head CT examination. The aims of this study were to investigate in vivo the levels of breast exposure to scatter radiation in head CT examination and the dependence of breast exposure upon body constitution. We tried to estimate the efficacy of external lead shielding as a mean of breast dose reduction, and to determine how much radiation reaches the organ from outside, in comparison to radiation load caused through internal scatter. We conclude that, although the level of breast radiation exposure during head CT examinations is generally low, shielding of the breasts with lead apron will further reduce the doses. However is the effect of shielding limited

  15. Usefulness of SPECT/CT in the Diagnosis of Intrathoracic Goiter versus Metastases From Cancer of the Breast

    DEFF Research Database (Denmark)

    Dümcke, Christine Elisabeth; Madsen, Jan Lysgård

    2007-01-01

    A 77-year-old woman was referred because of local reoccurrence of cancer of the breast. Chest x-ray showed a mediastinal tumor with dislocation of the trachea to the right. A Tc-99m pertechnetate scan showed irregular tracer uptake in an enlarged left lobe of the thyroid gland. Ultrasound confirmed...... the diagnosis of a nodular goiter with intrathoracic growth of the left lobe. A SPECT/CT scan of the mediastinum clearly showed that the soft tissue tumor was the left lobe of the thyroid gland, and not lymphatic metastases....

  16. An angle-dependent estimation of CT x-ray spectrum from rotational transmission measurements

    International Nuclear Information System (INIS)

    Lin, Yuan; Samei, Ehsan; Ramirez-Giraldo, Juan Carlos; Gauthier, Daniel J.; Stierstorfer, Karl

    2014-01-01

    Purpose: Computed tomography (CT) performance as well as dose and image quality is directly affected by the x-ray spectrum. However, the current assessment approaches of the CT x-ray spectrum require costly measurement equipment and complicated operational procedures, and are often limited to the spectrum corresponding to the center of rotation. In order to address these limitations, the authors propose an angle-dependent estimation technique, where the incident spectra across a wide range of angular trajectories can be estimated accurately with only a single phantom and a single axial scan in the absence of the knowledge of the bowtie filter. Methods: The proposed technique uses a uniform cylindrical phantom, made of ultra-high-molecular-weight polyethylene and positioned in an off-centered geometry. The projection data acquired with an axial scan have a twofold purpose. First, they serve as a reflection of the transmission measurements across different angular trajectories. Second, they are used to reconstruct the cross sectional image of the phantom, which is then utilized to compute the intersection length of each transmission measurement. With each CT detector element recording a range of transmission measurements for a single angular trajectory, the spectrum is estimated for that trajectory. A data conditioning procedure is used to combine information from hundreds of collected transmission measurements to accelerate the estimation speed, to reduce noise, and to improve estimation stability. The proposed spectral estimation technique was validated experimentally using a clinical scanner (Somatom Definition Flash, Siemens Healthcare, Germany) with spectra provided by the manufacturer serving as the comparison standard. Results obtained with the proposed technique were compared against those obtained from a second conventional transmission measurement technique with two materials (i.e., Cu and Al). After validation, the proposed technique was applied to measure

  17. Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT

    International Nuclear Information System (INIS)

    Angel, Erin; Yaghmai, Nazanin; Jude, Cecilia Matilda; DeMarco, John J; Cagnon, Christopher H; Goldin, Jonathan G; McNitt-Gray, Michael F; Primak, Andrew N; McCollough, Cynthia H; Stevens, Donna M; Cody, Dianna D

    2009-01-01

    Tube current modulation was designed to reduce radiation dose in CT imaging while maintaining overall image quality. This study aims to develop a method for evaluating the effects of tube current modulation (TCM) on organ dose in CT exams of actual patient anatomy. This method was validated by simulating a TCM and a fixed tube current chest CT exam on 30 voxelized patient models and estimating the radiation dose to each patient's glandular breast tissue. This new method for estimating organ dose was compared with other conventional estimates of dose reduction. Thirty detailed voxelized models of patient anatomy were created based on image data from female patients who had previously undergone clinically indicated CT scans including the chest area. As an indicator of patient size, the perimeter of the patient was measured on the image containing at least one nipple using a semi-automated technique. The breasts were contoured on each image set by a radiologist and glandular tissue was semi-automatically segmented from this region. Previously validated Monte Carlo models of two multidetector CT scanners were used, taking into account details about the source spectra, filtration, collimation and geometry of the scanner. TCM data were obtained from each patient's clinical scan and factored into the model to simulate the effects of TCM. For each patient model, two exams were simulated: a fixed tube current chest CT and a tube current modulated chest CT. X-ray photons were transported through the anatomy of the voxelized patient models, and radiation dose was tallied in the glandular breast tissue. The resulting doses from the tube current modulated simulations were compared to the results obtained from simulations performed using a fixed mA value. The average radiation dose to the glandular breast tissue from a fixed tube current scan across all patient models was 19 mGy. The average reduction in breast dose using the tube current modulated scan was 17%. Results were

  18. Dynamic helical CT mammography of breast cancer

    International Nuclear Information System (INIS)

    Yamamoto, Akira; Fukushima, Hitoshi; Okamura, Ryuji; Nakamura, Yoshiaki; Morimoto, Taisuke; Urata, Yoji; Mukaihara, Sumio; Hayakawa, Katsumi

    2006-01-01

    The purpose of this study was to determine whether dynamic helical computed tomography (CT)-mammography could assist in selecting the most appropriate surgical method in women with breast cancer. Preoperative contrast-enhanced helical CT scanning of the breast was performed on 133 female patients with suspicion of breast cancer at the same time as clinical, mammographic, and/or ultrasonographic examinations. The patients were scanned in the prone position with a specially designed CT-compatible device. A helical scan was made with rapid intravenous bolus injection (3 ml/s) of 100 ml of iodine contrast material. Three-dimensional maximum intensity projection (MIP) images were reconstructed, and CT findings were correlated with surgical and histopathological findings. Histopathological analysis revealed 84 malignant lesions and seven benign lesions. The sensitivity, specificity, and accuracy levels of the CT scanning were 94.6%, 58.6%, and 78.9%. Helical scanning alone revealed additional contralateral carcinomas in three of four patients and additional ipsilateral carcinomas in three of five patients. However, the technique gave false-positive readings in 24 patients. The preoperative CT-mammogram altered the surgical method in six patients. Dynamic helical CT-mammography in the prone position may be one of the choices of adjunct imaging in patients with suspected breast cancer scheduled for surgery. (author)

  19. 50 μm pixel pitch wafer-scale CMOS active pixel sensor x-ray detector for digital breast tomosynthesis.

    Science.gov (United States)

    Zhao, C; Konstantinidis, A C; Zheng, Y; Anaxagoras, T; Speller, R D; Kanicki, J

    2015-12-07

    Wafer-scale CMOS active pixel sensors (APSs) have been developed recently for x-ray imaging applications. The small pixel pitch and low noise are very promising properties for medical imaging applications such as digital breast tomosynthesis (DBT). In this work, we evaluated experimentally and through modeling the imaging properties of a 50 μm pixel pitch CMOS APS x-ray detector named DynAMITe (Dynamic Range Adjustable for Medical Imaging Technology). A modified cascaded system model was developed for CMOS APS x-ray detectors by taking into account the device nonlinear signal and noise properties. The imaging properties such as modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE) were extracted from both measurements and the nonlinear cascaded system analysis. The results show that the DynAMITe x-ray detector achieves a high spatial resolution of 10 mm(-1) and a DQE of around 0.5 at spatial frequencies  CMOS APS x-ray detector, image aquisition geometry and image reconstruction techniques should be considered.

  20. X-ray CT monitoring of iceball growth and thermal distribution during cryosurgery

    International Nuclear Information System (INIS)

    Sandison, G.A.; Loye, M.P.; Rewcastle, J.C.; Hahn, L.J.; Saliken, J.C.; McKinnon, J.G.; Donnelly, B.J.

    1998-01-01

    X-ray CT is able to image the internal architecture of frozen tissue. Phantoms of distilled water, a saline-gelatin mixture, lard and a calf liver-gelatin suspension cooled by a plastic tube acting as a long liquid nitrogen cryoprobe were used to study the relationship between Hounsfield unit (HU) values and temperature. There is a signature change in HU value from unfrozen to completely frozen tissue. No discernible relation exists between temperature in a completely frozen tissue and its HU value for the temperature range achieved with commercial cryoprobes. However, such a relation does exist in the typically narrow region of phase change and it is this change in HU value that is the parameter of concern for quantitative monitoring of the freezing process. Calibration of temperature against change in HU value allows a limited set of isotherms to be generated in the phase change region for direct monitoring of iceball growth. The phase change temperature range, mid-phase change temperature and the absolute value of HU change from completely frozen to unfrozen tissue are shown to be sensitive to the medium. Modelling of the temperature distribution within the region of completely frozen phantom using the infinite cylinder solution to the Fourier heat equation allows the temperature history of the phantom to be predicted. A set of isotherms, generated using a combination of thermal modelling and calibrated HU values demonstrates the feasibility of routine x-ray CT assisted cryotherapy. Isotherm overlay will be a major aid to the cryosurgeon who adopts a fixed target temperature as the temperature below which there is a certainty of ablation of the diseased tissue. (author)

  1. X-ray CT monitoring of iceball growth and thermal distribution during cryosurgery

    Energy Technology Data Exchange (ETDEWEB)

    Sandison, G.A.; Loye, M.P.; Rewcastle, J.C. [Departments of Oncology and Medical Physics, Tom Baker Cancer Centre, Calgary, Canada T2N 4N2 (Canada); Department of Physics and Astronomy, University of Calgary, Calgary, Canada T2N 2N4 (Canada); Hahn, L.J. [Department of Physics and Astronomy, University of Calgary, Calgary, Canada T2N 2N4 (Canada); Department of Diagnostic Imaging, Foothills Hospital, Calgary, Canada T2N 2T7 (Canada); Saliken, J.C. [Department of Diagnostic Imaging, Foothills Hospital, Calgary, Canada T2N 2T7 (Canada); Department of Surgery, Tom Baker Cancer Centre, Calgary, Canada T2N 4N2 (Canada); McKinnon, J.G. [Department of Surgery, Foothills Hospital, Calgary, Canada T2N 2T7 (Canada); Donnelly, B.J. [Department of Surgery, Tom Baker Cancer Centre, Calgary, Canada T2N 4N2 (Canada)

    1998-11-01

    X-ray CT is able to image the internal architecture of frozen tissue. Phantoms of distilled water, a saline-gelatin mixture, lard and a calf liver-gelatin suspension cooled by a plastic tube acting as a long liquid nitrogen cryoprobe were used to study the relationship between Hounsfield unit (HU) values and temperature. There is a signature change in HU value from unfrozen to completely frozen tissue. No discernible relation exists between temperature in a completely frozen tissue and its HU value for the temperature range achieved with commercial cryoprobes. However, such a relation does exist in the typically narrow region of phase change and it is this change in HU value that is the parameter of concern for quantitative monitoring of the freezing process. Calibration of temperature against change in HU value allows a limited set of isotherms to be generated in the phase change region for direct monitoring of iceball growth. The phase change temperature range, mid-phase change temperature and the absolute value of HU change from completely frozen to unfrozen tissue are shown to be sensitive to the medium. Modelling of the temperature distribution within the region of completely frozen phantom using the infinite cylinder solution to the Fourier heat equation allows the temperature history of the phantom to be predicted. A set of isotherms, generated using a combination of thermal modelling and calibrated HU values demonstrates the feasibility of routine x-ray CT assisted cryotherapy. Isotherm overlay will be a major aid to the cryosurgeon who adopts a fixed target temperature as the temperature below which there is a certainty of ablation of the diseased tissue. (author)

  2. A nanotube-based field emission x-ray source for microcomputed tomography

    International Nuclear Information System (INIS)

    Zhang, J.; Cheng, Y.; Lee, Y.Z.; Gao, B.; Qiu, Q.; Lin, W.L.; Lalush, D.; Lu, J.P.; Zhou, O.

    2005-01-01

    Microcomputed tomography (micro-CT) is a noninvasive imaging tool commonly used to probe the internal structures of small animals for biomedical research and for the inspection of microelectronics. Here we report the development of a micro-CT scanner with a carbon nanotube- (CNT-) based microfocus x-ray source. The performance of the CNT x-ray source and the imaging capability of the micro-CT scanner were characterized

  3. Assessment of Brain absorbed X-ray dose during CT- Scan using ImPACT software in Tehran Univeristy hospitals

    Directory of Open Access Journals (Sweden)

    Khalilpour M

    2009-07-01

    Full Text Available "n Normal 0 false false false EN-US X-NONE AR-SA MicrosoftInternetExplorer4 st1":*{behavior:url(#ieooui } /* Style Definitions */ table.MsoNormalTable {mso-style-name:"Table Normal"; mso-tstyle-rowband-size:0; mso-tstyle-colband-size:0; mso-style-noshow:yes; mso-style-priority:99; mso-style-qformat:yes; mso-style-parent:""; mso-padding-alt:0cm 5.4pt 0cm 5.4pt; mso-para-margin:0cm; mso-para-margin-bottom:.0001pt; mso-pagination:widow-orphan; font-size:11.0pt; font-family:"Calibri","sans-serif"; mso-ascii-font-family:Calibri; mso-ascii-theme-font:minor-latin; mso-fareast-font-family:"Times New Roman"; mso-fareast-theme-font:minor-fareast; mso-hansi-font-family:Calibri; mso-hansi-theme-font:minor-latin; mso-bidi-font-family:Arial; mso-bidi-theme-font:minor-bidi;} Background: CT scan was first introduced into clinical practice in 1972, and since then has grown into one of the predominant diagnostic procedures. In 1998, the UK National Radiological Protection Board reported that 20% of the national collective dose from medical X-ray examinations derived from CT-scans, although it represented only 2% of all X- ray examinations the aim of this study was to determine the X-ray dosage received by patients in brain CT scan."n"n Methods: In this work, we have estimated patient dose arising from CT examination of brain in five hospitals in Tehran. Organ and effective doses were estimated for 150 patients who underwent CT examination of brain. "ImPACT" version 0.99v was used to estimate organ and effective dose. Brain examinations were performed with fixed Kvp, mAs and T (slice thickness for each scanner. "n"n Results: Patients, who were scanned by CT of emam Khomeini center (Toshiba Xvision /EX Scanner, received maximum organ dose (brain and minimum organ dose was delivered to patients who were scanned by CT of amir alam center (Toshiba Xvision /EX Scanner. Maximum effective dose was 1.7 mSv acquired in this study for emam Khomeini haspital, smaller than

  4. Microcalcification detectability using a bench-top prototype photon-counting breast CT based on a Si strip detector.

    Science.gov (United States)

    Cho, Hyo-Min; Ding, Huanjun; Barber, William C; Iwanczyk, Jan S; Molloi, Sabee

    2015-07-01

    To investigate the feasibility of detecting breast microcalcification (μCa) with a dedicated breast computed tomography (CT) system based on energy-resolved photon-counting silicon (Si) strip detectors. The proposed photon-counting breast CT system and a bench-top prototype photon-counting breast CT system were simulated using a simulation package written in matlab to determine the smallest detectable μCa. A 14 cm diameter cylindrical phantom made of breast tissue with 20% glandularity was used to simulate an average-sized breast. Five different size groups of calcium carbonate grains, from 100 to 180 μm in diameter, were simulated inside of the cylindrical phantom. The images were acquired with a mean glandular dose (MGD) in the range of 0.7-8 mGy. A total of 400 images was used to perform a reader study. Another simulation study was performed using a 1.6 cm diameter cylindrical phantom to validate the experimental results from a bench-top prototype breast CT system. In the experimental study, a bench-top prototype CT system was constructed using a tungsten anode x-ray source and a single line 256-pixels Si strip photon-counting detector with a pixel pitch of 100 μm. Calcium carbonate grains, with diameter in the range of 105-215 μm, were embedded in a cylindrical plastic resin phantom to simulate μCas. The physical phantoms were imaged at 65 kVp with an entrance exposure in the range of 0.6-8 mGy. A total of 500 images was used to perform another reader study. The images were displayed in random order to three blinded observers, who were asked to give a 4-point confidence rating on each image regarding the presence of μCa. The μCa detectability for each image was evaluated by using the average area under the receiver operating characteristic curve (AUC) across the readers. The simulation results using a 14 cm diameter breast phantom showed that the proposed photon-counting breast CT system can achieve high detection accuracy with an average AUC greater

  5. Automatic quantification of mammary glands on non-contrast x-ray CT by using a novel segmentation approach

    Science.gov (United States)

    Zhou, Xiangrong; Kano, Takuya; Cai, Yunliang; Li, Shuo; Zhou, Xinxin; Hara, Takeshi; Yokoyama, Ryujiro; Fujita, Hiroshi

    2016-03-01

    This paper describes a brand new automatic segmentation method for quantifying volume and density of mammary gland regions on non-contrast CT images. The proposed method uses two processing steps: (1) breast region localization, and (2) breast region decomposition to accomplish a robust mammary gland segmentation task on CT images. The first step detects two minimum bounding boxes of left and right breast regions, respectively, based on a machine-learning approach that adapts to a large variance of the breast appearances on different age levels. The second step divides the whole breast region in each side into mammary gland, fat tissue, and other regions by using spectral clustering technique that focuses on intra-region similarities of each patient and aims to overcome the image variance caused by different scan-parameters. The whole approach is designed as a simple structure with very minimum number of parameters to gain a superior robustness and computational efficiency for real clinical setting. We applied this approach to a dataset of 300 CT scans, which are sampled with the equal number from 30 to 50 years-old-women. Comparing to human annotations, the proposed approach can measure volume and quantify distributions of the CT numbers of mammary gland regions successfully. The experimental results demonstrated that the proposed approach achieves results consistent with manual annotations. Through our proposed framework, an efficient and effective low cost clinical screening scheme may be easily implemented to predict breast cancer risk, especially on those already acquired scans.

  6. Measurement of erosion of stainless steel by molten lead-free solder using micro-focus x-ray CT system

    International Nuclear Information System (INIS)

    Nishikawa, Hiroshi; Takemoto, Tadashi; Kang, Songai

    2009-01-01

    The severe erosion damage, which is caused by a molten lead-free solder, of wave solder equipment made into stainless steel has been encountered in operation. Then, the higher maintenance frequency and reduced life time of wave solder machine component is a serious issue in a manufacturing process. In this study, the evaluation method of erosion of stainless steel by molten lead-free solders was investigated using micro-focus X-ray systems for fluoroscopic and computed tomography (CT). As a result, it was found that the fluoroscopic image could truly reconstruct the cross-shape of the stainless steel sample after immersion test without destruction. In the case of X-ray systems for fluoroscopic and CT used in this study, three-dimensional data can be obtained. Therefore, it was possible to easily check the whole picture of the test sample after immersion test and to decide the maximum erosion depth of test sample. (author)

  7. Analysis of Patients' X-ray Exposure in 146 Percutaneous Radiologic Gastrostomies.

    Science.gov (United States)

    Petersen, Tim-Ole; Reinhardt, Martin; Fuchs, Jochen; Gosch, Dieter; Surov, Alexey; Stumpp, Patrick; Kahn, Thomas; Moche, Michael

    2017-09-01

    Purpose  Analysis of patient´s X-ray exposure during percutaneous radiologic gastrostomies (PRG) in a larger population. Materials and Methods  Data of primary successful PRG-procedures, performed between 2004 and 2015 in 146 patients, were analyzed regarding the exposition to X-ray. Dose-area-product (DAP), dose-length-product (DLP) respectively, and fluoroscopy time (FT) were correlated with the used x-ray systems (Flatpanel Detector (FD) vs. Image Itensifier (BV)) and the necessity for periprocedural placement of a nasogastric tube. Additionally, the effective X-ray dose for PRG placement using fluoroscopy (DL), computed tomography (CT), and cone beam CT (CBCT) was estimated using a conversion factor. Results  The median DFP of PRG-placements under fluoroscopy was 163 cGy*cm 2 (flat panel detector systems: 155 cGy*cm 2 ; X-ray image intensifier: 175 cGy*cm 2 ). The median DLZ was 2.2 min. Intraprocedural placement of a naso- or orogastric probe (n = 68) resulted in a significant prolongation of the median DLZ to 2.5 min versus 2 min in patients with an already existing probe. In addition, dose values were analyzed in smaller samples of patients in which the PRG was placed under CBCT (n = 7, median DFP = 2635 cGy*cm 2 ), or using CT (n = 4, median DLP = 657 mGy*cm). Estimates of the median DFP and DLP showed effective doses of 0.3 mSv for DL-assisted placements (flat panel detector 0.3 mSv, X-ray image converter 0.4 mSv), 7.9 mSv using a CBCT - flat detector, and 9.9 mSv using CT. This corresponds to a factor 26 of DL versus CBCT, or a factor 33 of DL versus CT. Conclusion  In order to minimize X-ray exposure during PRG-procedures for patients and staff, fluoroscopically-guided interventions should employ flat detector systems with short transmittance sequences in low dose mode and with slow image frequency. Series recordings can be dispensed with. The intraprocedural placement of a naso- or orogastric probe

  8. Bi-directional x-ray phase-contrast mammography.

    Directory of Open Access Journals (Sweden)

    Kai Scherer

    Full Text Available Phase-contrast x-ray imaging is a promising improvement of conventional absorption-based mammography for early tumor detection. This potential has been demonstrated recently, utilizing structured gratings to obtain differential phase and dark-field scattering images. However, the inherently anisotropic imaging sensitivity of the proposed mono-directional approach yields only insufficient diagnostic information, and has low diagnostic sensitivity to highly oriented structures. To overcome these limitations, we present a two-directional x-ray phase-contrast mammography approach and demonstrate its advantages by applying it to a freshly dissected, cancerous mastectomy breast specimen. We illustrate that the two-directional scanning procedure overcomes the insufficient diagnostic value of a single scan, and reliably detects tumor structures, independently from their orientation within the breast. Our results indicate the indispensable diagnostic necessity and benefit of a multi-directional approach for x-ray phase-contrast mammography.

  9. Three dimensional analysis of coelacanth body structure by computer graphics and X-ray CT images

    International Nuclear Information System (INIS)

    Suzuki, Naoki; Hamada, Takashi.

    1990-01-01

    Three dimensional imaging processes were applied for the structural and functional analyses of the modern coelacanth (Latimeria chalumnae). Visualization of the obtained images is performed with computer graphics on the basis of serial images by an X-ray CT scanning method. Reconstruction of three dimensional images of the body structure of coelacanth using the volume rendering and surface rendering methods provides us various information about external and internal shapes of this exquisite fish. (author)

  10. Phase-contrast tomographic imaging using an X-ray interferometer

    Energy Technology Data Exchange (ETDEWEB)

    Momose, A. [Hitachi Ltd, Advanced Research Lab., Saitama (Japan); Takeda, T.; Itai, Y. [Univ. of Tsukuba, Inst. of Clinical Medicine, Ibaraki (Japan); Yoneyama, A. [Hitachi Ltd, Central Resarch Lab., Tokyo (Japan); Hirano, K. [High Energy Accelerator Research Organization, Inst. of Materials Structure Science, Ibaraki (Japan)

    1998-05-01

    Apparatus for phase-contrast X-ray computed tomography using a monolithic X-ray interferometer is presented with some observational results for human breast tissues. Structures characteristic of the tissues were revealed in the phase-contrast tomograms. The procedure of image analysis consists of phase retrieval from X-ray interference patterns and tomographic image reconstruction from the retrieved phase shift. Next, feasibility of phase-contrast imaging using a two-crystal X-ray interferometer was studied aiming at in vivo observation in the future. In a preliminary study, the two-crystal X-ray interferometer was capable of generating fringes of 70% visibility using synchrotron X-rays. 35 refs.

  11. Phase-contrast tomographic imaging using an X-ray interferometer

    International Nuclear Information System (INIS)

    Momose, A.; Takeda, T.; Itai, Y.; Yoneyama, A.; Hirano, K.

    1998-01-01

    Apparatus for phase-contrast X-ray computed tomography using a monolithic X-ray interferometer is presented with some observational results for human breast tissues. Structures characteristic of the tissues were revealed in the phase-contrast tomograms. The procedure of image analysis consists of phase retrieval from X-ray interference patterns and tomographic image reconstruction from the retrieved phase shift. Next, feasibility of phase-contrast imaging using a two-crystal X-ray interferometer was studied aiming at in vivo observation in the future. In a preliminary study, the two-crystal X-ray interferometer was capable of generating fringes of 70% visibility using synchrotron X-rays

  12. Photon activated therapy (PAT) using monochromatic synchrotron X-rays and iron oxide nanoparticles in a mouse tumor model: feasibility study of PAT for the treatment of superficial malignancy.

    Science.gov (United States)

    Choi, Gi-Hwan; Seo, Seung-Jun; Kim, Ki-Hong; Kim, Hong-Tae; Park, Sung-Hwan; Lim, Jae-Hong; Kim, Jong-Ki

    2012-10-31

    X-rays are known to interact with metallic nanoparticles, producing photoelectric species as radiosensitizing effects, and have been exploited in vivo mainly with gold nanoparticles. The purpose of this study was to investigate the potential of sensitizing effect of iron oxide nanoparticles for photon activated therapy. X-rays photon activated therapy (PAT) was studied by treating CT26 tumor cells and CT26 tumor-bearing mice loaded with 13-nm diameter FeO NP, and irradiating them at 7.1 keV near the Fe K-edge using synchrotron x-rays radiation. Survival of cells was determined by MTT assay, and tumor regression assay was performed for in vivo model experiment. The results of PAT treated groups were compared with x-rays alone control groups. A more significant reduction in viability and damage was observed in the FeO NP-treated irradiated cells, compared to the radiation alone group (p X-rays. Since 7.1 keV X-rays is attenuated very sharply in the tissue, FeO NP-PAT may have promise as a potent treatment option for superficial malignancies in the skin, like chest wall recurrence of breast cancer.

  13. Compact energy dispersive X-ray microdiffractometer for diagnosis of neoplastic tissues

    Science.gov (United States)

    Sosa, C.; Malezan, A.; Poletti, M. E.; Perez, R. D.

    2017-08-01

    An energy dispersive X-ray microdiffractometer with capillary optics has been developed for characterizing breast cancer. The employment of low divergence capillary optics helps to reduce the setup size to a few centimeters, while providing a lateral spatial resolution of 100 μm. The system angular calibration and momentum transfer resolution were assessed by a detailed study of a polycrystalline reference material. The performance of the system was tested by means of the analysis of tissue-equivalent samples previously characterized by conventional X-ray diffraction. In addition, a simplified correction model for an appropriate comparison of the diffraction spectra was developed and validated. Finally, the system was employed to evaluate normal and neoplastic human breast samples, in order to determine their X-ray scatter signatures. The initial results indicate that the use of this compact energy dispersive X-ray microdiffractometer combined with a simplified correction procedure is able to provide additional information to breast cancer diagnosis.

  14. Eigenvector decomposition of full-spectrum x-ray computed tomography.

    Science.gov (United States)

    Gonzales, Brian J; Lalush, David S

    2012-03-07

    Energy-discriminated x-ray computed tomography (CT) data were projected onto a set of basis functions to suppress the noise in filtered back-projection (FBP) reconstructions. The x-ray CT data were acquired using a novel x-ray system which incorporated a single-pixel photon-counting x-ray detector to measure the x-ray spectrum for each projection ray. A matrix of the spectral response of different materials was decomposed using eigenvalue decomposition to form the basis functions. Projection of FBP onto basis functions created a de facto image segmentation of multiple contrast agents. Final reconstructions showed significant noise suppression while preserving important energy-axis data. The noise suppression was demonstrated by a marked improvement in the signal-to-noise ratio (SNR) along the energy axis for multiple regions of interest in the reconstructed images. Basis functions used on a more coarsely sampled energy axis still showed an improved SNR. We conclude that the noise-resolution trade off along the energy axis was significantly improved using the eigenvalue decomposition basis functions.

  15. A phantom-based evaluation of three commercially available patient organ shields for computed tomography X-ray examinations in diagnostic radiology

    International Nuclear Information System (INIS)

    Huggett, J.; Mukonoweshuro, W.; Loader, R.

    2013-01-01

    Three commercially available in-plane patient organ shields (barium eye, bismuth eye and bismuth breast) for computed tomography (CT) examinations were evaluated to determine their effectiveness for dose reduction. Absorbed doses were measured using metal oxide semiconductor field effect transistor dosemeters fastened to a Kyoto CT Torso phantom. Resultant images were visually compared with those minus shielding by an experienced radiologist. Approximate dose reductions of 21, 38 and 50 % were achieved by the barium eye, bismuth eye and bismuth breast shields, respectively, at a cost of increased image noise and streak artefacts. Shielded images produced varied levels of image artefact, particularly those resulting from the eye shields. Measured dose reductions were not consistent with the potential dose savings stated by the manufacturers of the shields. When evaluating the breast shield, similar dose reduction was achieved without shield-induced artefact by simply reducing the X-ray tube current. (authors)

  16. 3d Finite Element Modelling of Non-Crimp Fabric Based Fibre Composite Based on X-Ray Ct Data

    DEFF Research Database (Denmark)

    Jespersen, Kristine Munk; Asp, Leif; Mikkelsen, Lars Pilgaard

    2017-01-01

    initiation and progression in the material. In the current study, the real bundle structure inside a non-crimp fabric based fibre composite is extracted from 3D X-ray CT images and imported into ABAQUS for numerical modelling.The local stress concentrations when loaded in tension caused by the fibre bundle...

  17. Phantom studies on the artifacts of barium on 18F-FDG DHC/CT images induced by X-ray attenuation correction

    International Nuclear Information System (INIS)

    Wang Wei; Zhu Jiarui; Wang Xinqiang; Zhao Wenrui; Chuan Ling; Xu Genxiang; Gao Chunhua; Fang Tingzheng

    2007-01-01

    Objective: Attenuation correction (AC) based on X-ray transmission map may result in false positive readings or artifacts on PET images, some of them due to the internal residue of high density contrast media used in diagnostic X-ray imaging. The aim of this study was to experimentally estimate the impacts of different concentrations and volumes of barium contrast on X-CT AC (CTAC) for dual-head coincidence (DHC/CT) images. Methods: A cylindrical phantom containing 18 F solution (3.7 kBq/ml), in which plastic fingertips enclosed with different concentrations (0.05, 0.1, 0.2, 0.3, 0.4, 0.5, 1.0, 2.0, 3.0 kg/L)and volumes(0.5, 1.0, 1.5, 2.0, 2.5 ml) of BaSO 4 contrast media were inserted, was used to modulate routine 18 F-fluorodeoxyglucose (FDG) imaging study on a DHC/CT scanner (GE Discovery VH). Sequential 18 F emission and X-CT transmission acquisitions followed by data processing and reconstruction were carried out in clinical settings. For comparison, both visual and quantitative analyses were performed on CTAC and non-AC (NAC) images of the phantom. Results: In NAC images, the radioactivity distribution within the whole phantom was non-uniform with lower counts in the center; the plastic fingertips were all seen as 'cold spots' with much lower counts in the contrast region than in their surrounding areas. On the contrary, in CTAC images, the radioactivity distribution within the whole phantom was almost uniform; while most plastic fingertips with media concentration ≥0.1 kg/L and volume >0.5 ml were all depicted as 'hot spots' with higher counts than in surrounding areas. Conclusions: Barium contrast with relative high concentration or large volume can induce artifacts on CTAC DHC/CT images. In clinical setting, proper interpretation of CTAC DHC/CT images should refer to NAC DHC/CT images to exclude any artifacts related to the contrast media residues. (authors)

  18. X-ray CT for quantitative food microstructure engineering: The apple case

    International Nuclear Information System (INIS)

    Herremans, Els; Verboven, Pieter; Defraeye, Thijs; Rogge, Seppe; Ho, Quang Tri; Hertog, Maarten L.A.T.M.; Verlinden, Bert E.; Bongaers, Evi; Wevers, Martine; Nicolai, Bart M.

    2014-01-01

    Highlights: • Diffusivity of O 2 and CO 2 in Braeburn tissue were computed from X-ray micro-CT data. • Diffusivities were different for the distinct stages of ‘Braeburn’ browning disorder. • Microstructural gas transport constants were implemented in models of intact fruit. • We simulated internal O 2 concentrations for longterm storage treatments of Braeburn. • This engineering approach can optimize internal quality of apple during CA storage. - Abstract: Apple fruit is a major crop that can be supplied year-round due to low temperature storage in a controlled atmosphere with a reduced oxygen concentration and an increased carbon dioxide concentration. The low temperature and dedicated gas concentration levels are designed to provide optimal conditions that prevent ripening while maintaining the fundamental respiratory metabolism necessary for energy supply in the cells that ensures cell and tissue integrity during storage of the fruit. If the concentration of oxygen is too low or that of carbon dioxide too high, a fermentation metabolism is induced that causes the production of off-flavours, results in insufficient energy supply, leading to cell collapse and consequent tissue browning and cavity formation. The microstructural arrangement of cells and intercellular spaces in the apple create specific pathways for transport of the respiratory gasses oxygen and carbon dioxide. We used X-ray CT to characterise the changes in the microstructure of ‘Braeburn’ apple during the development of internal storage disorders. Multiscale modeling was applied to understand the changes in oxygen and carbon dioxide concentrations and respiration and fermentation rates in the apple during the disorder development in controlled atmosphere storage of ‘Braeburn’ apple fruit. The 3D microstructure geometries of healthy, brown tissue and tissue with cavities were created to solve the micro-scale gas-exchange model for O 2 and CO 2 using the finite volume method

  19. X-ray CT for quantitative food microstructure engineering: The apple case

    Energy Technology Data Exchange (ETDEWEB)

    Herremans, Els, E-mail: els.herremans@biw.kuleuven.be [BIOSYST-MeBioS, KU Leuven, Willem de Croylaan 42, 3001 Leuven (Belgium); Verboven, Pieter; Defraeye, Thijs; Rogge, Seppe; Ho, Quang Tri; Hertog, Maarten L.A.T.M. [BIOSYST-MeBioS, KU Leuven, Willem de Croylaan 42, 3001 Leuven (Belgium); Verlinden, Bert E. [VCBT, Flanders Centre of Postharvest Technology, Willem de Croylaan 42, 3001 Leuven (Belgium); Bongaers, Evi [Bruker microCT, Kartuizersweg 3B, 2550 Kontich (Belgium); Wevers, Martine [MTM, KU Leuven, Kasteelpark Arenberg 44, 3001 Leuven (Belgium); Nicolai, Bart M. [BIOSYST-MeBioS, KU Leuven, Willem de Croylaan 42, 3001 Leuven (Belgium); VCBT, Flanders Centre of Postharvest Technology, Willem de Croylaan 42, 3001 Leuven (Belgium)

    2014-04-01

    Highlights: • Diffusivity of O{sub 2} and CO{sub 2} in Braeburn tissue were computed from X-ray micro-CT data. • Diffusivities were different for the distinct stages of ‘Braeburn’ browning disorder. • Microstructural gas transport constants were implemented in models of intact fruit. • We simulated internal O{sub 2} concentrations for longterm storage treatments of Braeburn. • This engineering approach can optimize internal quality of apple during CA storage. - Abstract: Apple fruit is a major crop that can be supplied year-round due to low temperature storage in a controlled atmosphere with a reduced oxygen concentration and an increased carbon dioxide concentration. The low temperature and dedicated gas concentration levels are designed to provide optimal conditions that prevent ripening while maintaining the fundamental respiratory metabolism necessary for energy supply in the cells that ensures cell and tissue integrity during storage of the fruit. If the concentration of oxygen is too low or that of carbon dioxide too high, a fermentation metabolism is induced that causes the production of off-flavours, results in insufficient energy supply, leading to cell collapse and consequent tissue browning and cavity formation. The microstructural arrangement of cells and intercellular spaces in the apple create specific pathways for transport of the respiratory gasses oxygen and carbon dioxide. We used X-ray CT to characterise the changes in the microstructure of ‘Braeburn’ apple during the development of internal storage disorders. Multiscale modeling was applied to understand the changes in oxygen and carbon dioxide concentrations and respiration and fermentation rates in the apple during the disorder development in controlled atmosphere storage of ‘Braeburn’ apple fruit. The 3D microstructure geometries of healthy, brown tissue and tissue with cavities were created to solve the micro-scale gas-exchange model for O{sub 2} and CO{sub 2} using the

  20. In vivo quantification of plant starch reserves at micrometer resolution using X-ray microCT imaging and machine learning.

    Science.gov (United States)

    Earles, J Mason; Knipfer, Thorsten; Tixier, Aude; Orozco, Jessica; Reyes, Clarissa; Zwieniecki, Maciej A; Brodersen, Craig R; McElrone, Andrew J

    2018-03-08

    Starch is the primary energy storage molecule used by most terrestrial plants to fuel respiration and growth during periods of limited to no photosynthesis, and its depletion can drive plant mortality. Destructive techniques at coarse spatial scales exist to quantify starch, but these techniques face methodological challenges that can lead to uncertainty about the lability of tissue-specific starch pools and their role in plant survival. Here, we demonstrate how X-ray microcomputed tomography (microCT) and a machine learning algorithm can be coupled to quantify plant starch content in vivo, repeatedly and nondestructively over time in grapevine stems (Vitis spp.). Starch content estimated for xylem axial and ray parenchyma cells from microCT images was correlated strongly with enzymatically measured bulk-tissue starch concentration on the same stems. After validating our machine learning algorithm, we then characterized the spatial distribution of starch concentration in living stems at micrometer resolution, and identified starch depletion in live plants under experimental conditions designed to halt photosynthesis and starch production, initiating the drawdown of stored starch pools. Using X-ray microCT technology for in vivo starch monitoring should enable novel research directed at resolving the spatial and temporal patterns of starch accumulation and depletion in woody plant species. No claim to original US Government works New Phytologist © 2018 New Phytologist Trust.

  1. Monochromatic x-rays for low-dose digital mammography: preliminary results.

    Science.gov (United States)

    Yoon, Kwon-Ha; Kwon, Young Man; Choi, Byoung-Jung; Son, Hyun Hwa; Ryu, Cheol Woo; Chon, Kwon Su; Park, Seong Hoon; Juhng, Sun Kwan

    2012-12-01

    The feasibility of using monochromatic x-ray imaging generated from an x-ray tube and a multilayer reflector for digital mammography with a low radiation dose was examined. A multilayer mirror was designed to select the x-ray peak with an energy of 21.5 keV generated from an x-ray tube with a tungsten target and was fabricated by the ion-beam sputtering deposition system. Monochromatic x-ray images were obtained from an experimental digital mammography setup with a scanning stage. The performance of the system was evaluated using a breast phantom, a spectrometer, and a radiation dosimeter. We measured the contrast-to-noise ratio and performed the 10% modulation function test to determine image quality and resolution. The monochromatic beam from the multilayer reflector had a full-width-at-half-maximum of 0.9 keV at 21.5 keV, and the reflectivity was 0.70, which was 90% of the designed value. The polychromatic and monochromatic x-rays showed radiation doses of 0.497 and 0.0415 mGy, respectively. The monochromatic x-ray image shows fibers, calcifications, and masses more clearly than the polychromatic x-ray images do. The image contrast of the monochromatic x-rays was 1.85 times higher than that of the polychromatic x-rays. The experimental mammography setup had a spatial resolution of 7 lp/mm with both x-rays. Monochromatic x-rays generated using a multilayer mirror may be a useful diagnostic tool for breast examination by providing high contrast imaging with a low radiation dose.

  2. Rib necrosis after postoperative irradiation with 6MeV x-ray to breast cancer

    Energy Technology Data Exchange (ETDEWEB)

    Asakawa, H; Watarai, J; Otawa, H [Miyagi Prefectural Adult Disease Center, Natori (Japan)

    1975-04-01

    In order to examine quality of rays in radiation injury in the rib by high energy x-ray, radiation injury in the rib was roentgenologically followed up in the subjects that received postoperative irradiation only with 6 MeV x-ray to breast cancer. The subjects consisted of 79 patients with the age of 30 to 78 and were irradiated with 200 rads of 6 MeV x-ray 5 times a week delivered in 5 to 6 weeks postoperatively. Two fields were irradiated in a day and the total dose reached more than 5,000 rads. Roentgenologic follow up for more than 12 months revealed that rib necrosis occurred in 9 (11%) of 79 patients and that there was no relation to the age of patient. The necrosis was most likely to develop in the right second rib 10 to 23 months after the irradiation and the mean was 16 months. To the rib region where necrosis occurred, 1,880 to 2,230 ret were irradiated and the mean was 2,014 ret. There found no relation between the irradiation dose and occurrence of rib necrosis in the extent of 4,000 to 6,000 rads for tumor doses. Radiation injury in the lung was complicated in 8 (89%) of 9 patients with rib necrosis, indicating high incidence.

  3. Breast tissue classification using x-ray scattering measurements and multivariate data analysis

    Science.gov (United States)

    Ryan, Elaine A.; Farquharson, Michael J.

    2007-11-01

    This study utilized two radiation scatter interactions in order to differentiate malignant from non-malignant breast tissue. These two interactions were Compton scatter, used to measure the electron density of the tissues, and coherent scatter to obtain a measure of structure. Measurements of these parameters were made using a laboratory experimental set-up comprising an x-ray tube and HPGe detector. The breast tissue samples investigated comprise five different tissue classifications: adipose, malignancy, fibroadenoma, normal fibrous tissue and tissue that had undergone fibrocystic change. The coherent scatter spectra were analysed using a peak fitting routine, and a technique involving multivariate analysis was used to combine the peak fitted scatter profile spectra and the electron density values into a tissue classification model. The number of variables used in the model was refined by finding the sensitivity and specificity of each model and concentrating on differentiating between two tissues at a time. The best model that was formulated had a sensitivity of 54% and a specificity of 100%.

  4. Breast tissue classification using x-ray scattering measurements and multivariate data analysis

    Energy Technology Data Exchange (ETDEWEB)

    Ryan, Elaine A; Farquharson, Michael J [School of Allied Health Sciences, City University, Charterhouse Square, London EC1M 6PA (United Kingdom)

    2007-11-21

    This study utilized two radiation scatter interactions in order to differentiate malignant from non-malignant breast tissue. These two interactions were Compton scatter, used to measure the electron density of the tissues, and coherent scatter to obtain a measure of structure. Measurements of these parameters were made using a laboratory experimental set-up comprising an x-ray tube and HPGe detector. The breast tissue samples investigated comprise five different tissue classifications: adipose, malignancy, fibroadenoma, normal fibrous tissue and tissue that had undergone fibrocystic change. The coherent scatter spectra were analysed using a peak fitting routine, and a technique involving multivariate analysis was used to combine the peak fitted scatter profile spectra and the electron density values into a tissue classification model. The number of variables used in the model was refined by finding the sensitivity and specificity of each model and concentrating on differentiating between two tissues at a time. The best model that was formulated had a sensitivity of 54% and a specificity of 100%.

  5. The impact of the PET/CT in comparison with the same day contrast enhanced CT in breast cancer management

    International Nuclear Information System (INIS)

    Piperkova, E.; Raphael, B.; Altinyay, M.; Castellon, I.; Libes, R.; Abdel-Dayem, H.

    2006-01-01

    Full text: The aim of this study is to evaluate the impact of 18F-fluorodeoxyglucose FDG positron emission tomography with fused computerized tomography (PET/CT) in comparison with the same day contrast enhanced CT (CE-CT) in the breast cancer (BC) management. 68 studies in 48 BC patients, 8 for initial and 60 for restaging disease, after surgery, radiation/chemotherapy, for radiation therapy planning or evaluating treatment response were included. All patients underwent whole body PET/CT for diagnostic purposes followed by CE-CT diagnostic scans of selected body regions. PET/CT was performed approximately 90 minutes following 10-15 mCi of 18F-FDG on a GE Discovery PET/CT system. CT part acquired with low dose X-ray for localization and attenuation correction. The CE-CT was performed according to departmental protocol. Out of a total of 235 lesions in 68 PET/CT and CE-CT studies, 189 were concordant between PET/CT and CE-CT. However, there were 46 discordant lesions, which were verified by either follow-up or biopsy. PET/CT correctly identified 25 (True Positive-TP). CE-CT identified 2 TP lesions missed by PET/CT (False negative): one liver metastasis with necrosis, which is a known non-avid FDG, the other was a missed abdominal metastatic node, which did not change staging or treatment. PET/CT incorrectly identified 2 lesions (False Positive) while CE-CT incorrectly identified 17 FP. For evaluating treatment response in 40 follow up studies PET/CT reported complete response in 15, partial response in 11, stable disease in 2, progression in 5, and free of disease following surgery - in 7. The CE-CT described progression of the disease in 1 PET/CT true negative study and no progression in 2 TP PET/CT studies. In this study, PET/CT played more important role than CECT scans alone and provided an impact on the management of BC patients

  6. Thermoacoustic CT of the breast: pilot study observations

    Science.gov (United States)

    Kruger, Robert A.; Kiser, William L., Jr.; Romilly, A. P.; Scmidt, Phyllis

    2001-06-01

    In order to assess the potential clinical utility of using thermoacoustic computer tomography (TCT) to image the breast, we conducted a retrospective pilot study of 78 patients. We recruited patients in three age groups (50 years). The study population was further segregated into normal and suspicious based on the results of the previous x-ray mammography and ultrasound. Image quality was evaluated qualitatively by consensus of two trained mammographers using a 4-point scale. The appearance of normal anatomy, cysts, benign disease and cancer was noted. Patients were also asked to rate the comfort of the TCT exam and to indicate a personal preference for x-ray mammography or TCT. Analysis of the data indicated that TCT image quality was dependent upon both patient age and breast density, improving with both increasing breast density and decreasing patient age. Fibrocystic disease was well seen, cysts appearing as areas of low RF absorption. Fibroadenomas did not demonstrate contrast enhancement with the exception of one patient with associated atypical hyperplasia. Cancer displayed higher RF absorption than surrounding tissues in 4/7 patients in whom cancer was confirmed, including one patient with a 7-mm ductal carcinoma in situ (DCIS).

  7. Tissue Equivalent Phantom Design for Characterization of a Coherent Scatter X-ray Imaging System

    Science.gov (United States)

    Albanese, Kathryn Elizabeth

    Scatter in medical imaging is typically cast off as image-related noise that detracts from meaningful diagnosis. It is therefore typically rejected or removed from medical images. However, it has been found that every material, including cancerous tissue, has a unique X-ray coherent scatter signature that can be used to identify the material or tissue. Such scatter-based tissue-identification provides the advantage of locating and identifying particular materials over conventional anatomical imaging through X-ray radiography. A coded aperture X-ray coherent scatter spectral imaging system has been developed in our group to classify different tissue types based on their unique scatter signatures. Previous experiments using our prototype have demonstrated that the depth-resolved coherent scatter spectral imaging system (CACSSI) can discriminate healthy and cancerous tissue present in the path of a non-destructive x-ray beam. A key to the successful optimization of CACSSI as a clinical imaging method is to obtain anatomically accurate phantoms of the human body. This thesis describes the development and fabrication of 3D printed anatomical scatter phantoms of the breast and lung. The purpose of this work is to accurately model different breast geometries using a tissue equivalent phantom, and to classify these tissues in a coherent x-ray scatter imaging system. Tissue-equivalent anatomical phantoms were designed to assess the capability of the CACSSI system to classify different types of breast tissue (adipose, fibroglandular, malignant). These phantoms were 3D printed based on DICOM data obtained from CT scans of prone breasts. The phantoms were tested through comparison of measured scatter signatures with those of adipose and fibroglandular tissue from literature. Tumors in the phantom were modeled using a variety of biological tissue including actual surgically excised benign and malignant tissue specimens. Lung based phantoms have also been printed for future

  8. Dual-energy X-ray micro-CT imaging of hybrid Ni/Al open-cell foam

    International Nuclear Information System (INIS)

    Fíla, T.; Koudelka, P.; Zlámal, P.; Jiroušek, O.; Kumpová, I.; Vavřík, D.; Jung, A.

    2016-01-01

    In this paper, we employ dual-energy X-ray microfocus tomography (DECT) measurement to develop high-resolution finite element (FE) models that can be used for the numerical assessment of the deformation behaviour of hybrid Ni/Al foam subjected to both quasi-static and dynamic compressive loading. Cubic samples of hybrid Ni/Al open-cell foam with an edge length of [15]mm were investigated by the DECT measurement. The material was prepared using AlSi 7 Mg 0.3 aluminium foam with a mean pore size of [0.85]mm, coated with nanocrystalline nickel (crystallite size of approx. [50]nm) to form a surface layer with a theoretical thickness of [0.075]mm. CT imaging was carried out using state-of-the-art DSCT/DECT X-ray scanner developed at Centre of Excellence Telč. The device consists of a modular orthogonal assembly of two tube-detector imaging pairs, with an independent geometry setting and shared rotational stage mounted on a complex 16-axis CNC positioning system to enable unprecedented measurement variability for highly-detailed tomographical measurements. A sample of the metal foam was simultaneously irradiated using an XWT-240-SE reflection type X-ray tube and an XWT-160-TCHR transmission type X-ray tube. An enhanced dual-source sampling strategy was used for data acquisition. X-ray images were taken using XRD1622 large area GOS scintillator flat panel detectors with an active area of [410 × 410]mm and resolution [2048 × 2048]pixels. Tomographic scanning was performed in 1,200 projections with a 0.3 degree angular step to improve the accuracy of the generated models due to the very complex microstructure and high attenuation of the investigated material. Reconstructed data was processed using a dual-energy algorithm, and was used for the development of a 3D model and voxel model of the foam. The selected parameters of the models were compared with nominal parameters of the actual foam and showed good correlation

  9. Application of 3D X-ray CT data sets to finite element analysis

    International Nuclear Information System (INIS)

    Bossart, P.L.; Martz, H.E.; Brand, H.R.; Hollerbach, K.

    1995-01-01

    Finite Element Modeling (FEM) is becoming more important as industry drives toward concurrent engineering. A fundamental hindrance to fully exploiting the power of FEM is the human effort required to acquire complex part geometry, particularly as-built geometry, as a FEM mesh. Many Quantitative Non Destructive Evaluation (QNDE) techniques that produce three-dimensional (3D) data sets provide a substantial reduction in the effort required to apply FEM to as-built parts. This paper describes progress at LLNL on the application of 3D X-ray computed tomography (CT) data sets to more rapidly produce high-quality FEM meshes of complex, as-built geometries. Issues related to the volume segmentation of the 3D CT data as well as the use of this segmented data to tailor generic hexahedral FEM meshes to part specific geometries are discussed. The application of these techniques to FEM analysis in the medical field is reported here

  10. X-ray computed tomography imaging method which is immune to beam hardening effect

    International Nuclear Information System (INIS)

    Kanno, Ikuo; Uesaka, Akio; Nomiya, Seiichiro; Onabe, Hideaki

    2009-01-01

    For the easy treatment of cancers, early finding of them is an important theme of study. X-ray transmission measurement and computed tomography (CT) are powerful tools for finding cancers. The x-ray CT shows cross sectional view of human body and is able to detect small cancers such as 1 cm in diameter. The CT, however, gives very high dose exposure to human body: some 10 to 1000 times higher dose exposure than the chest radiography. It is not possible to have medical health check using CT frequently, in view of both individual and public accumulated dose exposures. The authors have been working on the reduction of dose exposure in x-ray transmission measurements in case of detecting iodine contrast media, which concentrates in cancers. In our method, energy information of x-rays is employed: in conventional x-ray transmission measurements, x-rays are measured as current and the energy of each x-ray is ignored. The numbers of x-ray events, φ 1 and φ 2 , of which energies are lower and higher than the one of iodine K-edge, respectively, are used for the estimation of iodine thickness in cancers. Moreover, high energy x-rays, which are not sensitive to the absorption by iodine, are cut by a filter made of higher atomic number material than iodine. We call this method filtered x-ray energy subtraction (FIX-ES) method. This FIX-ES method was shown twice as sensitive to iodine than current measurement method. With the choice of filter thickness, minimum dose exposure in FIX-ES is 30% of that when white x-rays are employed. In the study described above, we concentrated on the observation of cancer part. In this study, a cancer phantom in normal tissue is observed by FIX-ES method. The results are compared with the ones obtained by current measurement method. (author)

  11. Multi-modality PET-CT imaging of breast cancer in an animal model using nanoparticle x-ray contrast agent and 18F-FDG

    Science.gov (United States)

    Badea, C. T.; Ghaghada, K.; Espinosa, G.; Strong, L.; Annapragada, A.

    2011-03-01

    Multi-modality PET-CT imaging is playing an important role in the field of oncology. While PET imaging facilitates functional interrogation of tumor status, the use of CT imaging is primarily limited to anatomical reference. In an attempt to extract comprehensive information about tumor cells and its microenvironment, we used a nanoparticle xray contrast agent to image tumor vasculature and vessel 'leakiness' and 18F-FDG to investigate the metabolic status of tumor cells. In vivo PET/CT studies were performed in mice implanted with 4T1 mammary breast cancer cells.Early-phase micro-CT imaging enabled visualization 3D vascular architecture of the tumors whereas delayedphase micro-CT demonstrated highly permeable vessels as evident by nanoparticle accumulation within the tumor. Both imaging modalities demonstrated the presence of a necrotic core as indicated by a hypo-enhanced region in the center of the tumor. At early time-points, the CT-derived fractional blood volume did not correlate with 18F-FDG uptake. At delayed time-points, the tumor enhancement in 18F-FDG micro-PET images correlated with the delayed signal enhanced due to nanoparticle extravasation seen in CT images. The proposed hybrid imaging approach could be used to better understand tumor angiogenesis and to be the basis for monitoring and evaluating anti-angiogenic and nano-chemotherapies.

  12. Radiological response of ceramic and polymeric devices for breast brachytherapy

    International Nuclear Information System (INIS)

    Batista Nogueira, Luciana; Passos Ribeiro de Campos, Tarcisio

    2012-01-01

    In the present study, the radiological visibility of ceramic and polymeric devices implanted in breast phantom was investigated for future applications in brachytherapy. The main goal was to determine the radiological viability of ceramic and polymeric devices in vitro by performing simple radiological diagnostic methods such as conventional X-ray analysis and mammography due to its easy access to the population. The radiological response of ceramic and polymeric devices implanted in breast phantom was determined using conventional X-ray, mammography and CT analysis. - Highlights: ► Radiological visibility of ceramic and polymeric devices implanted in breast phantom. ► The barium incorporation in the seed improves the radiological contrast. ► Radiological monitoring shows the position, orientation and degradation of devices. ► Simple radiological methods such as X-ray and mammography were used for radiological monitoring.

  13. X-ray CT analysis of pore structure in sand

    Science.gov (United States)

    Mukunoki, Toshifumi; Miyata, Yoshihisa; Mikami, Kazuaki; Shiota, Erika

    2016-06-01

    The development of microfocused X-ray computed tomography (CT) devices enables digital imaging analysis at the pore scale. The applications of these devices are diverse in soil mechanics, geotechnical and geoenvironmental engineering, petroleum engineering, and agricultural engineering. In particular, the imaging of the pore space in porous media has contributed to numerical simulations for single-phase and multiphase flows or contaminant transport through the pore structure as three-dimensional image data. These obtained results are affected by the pore diameter; therefore, it is necessary to verify the image preprocessing for the image analysis and to validate the pore diameters obtained from the CT image data. Moreover, it is meaningful to produce the physical parameters in a representative element volume (REV) and significant to define the dimension of the REV. This paper describes the underlying method of image processing and analysis and discusses the physical properties of Toyoura sand for the verification of the image analysis based on the definition of the REV. On the basis of the obtained verification results, a pore-diameter analysis can be conducted and validated by a comparison with the experimental work and image analysis. The pore diameter is deduced from Young-Laplace's law and a water retention test for the drainage process. The results from previous study and perforated-pore diameter originally proposed in this study, called the voxel-percolation method (VPM), are compared in this paper. In addition, the limitations of the REV, the definition of the pore diameter, and the effectiveness of the VPM for an assessment of the pore diameter are discussed.

  14. X-ray Compton line scan tomography

    Energy Technology Data Exchange (ETDEWEB)

    Kupsch, Andreas; Lange, Axel; Jaenisch, Gerd-Ruediger [Bundesanstalt fuer Materialforschung und -pruefung (BAM), Berlin (Germany). Fachgruppe 8.5 - Mikro-ZfP; Hentschel, Manfred P. [Technische Univ. Berlin (Germany); Kardjilov, Nikolay; Markoetter, Henning; Hilger, Andre; Manke, Ingo [Helmholtz-Zentrum Berlin (HZB) (Germany); Toetzke, Christian [Potsdam Univ. (Germany)

    2015-07-01

    The potentials of incoherent X-ray scattering (Compton) computed tomography (CT) are investigated. The imaging of materials of very different atomic number or density at once is generally a perpetual challenge for X-ray tomography or radiography. In a basic laboratory set-up for simultaneous perpendicular Compton scattering and direct beam attenuation tomography are conducted by single channel photon counting line scans. This results in asymmetric distortions of the projection profiles of the scattering CT data set. In a first approach, corrections of Compton scattering data by taking advantage of rotational symmetry yield tomograms without major geometric artefacts. A cylindrical sample composed of PE, PA, PVC, glass and wood demonstrates similar Compton contrast for all the substances, while the conventional absorption tomogram only reveals the two high order materials. Comparison to neutron tomography reveals astonishing similarities except for the glass component (without hydrogen). Therefore, Compton CT offers the potential to replace neutron tomography, which requires much more efforts.

  15. Ultrasound detection of pneumothorax compared with chest X-ray and computed tomography scan.

    Science.gov (United States)

    Nagarsheth, Khanjan; Kurek, Stanley

    2011-04-01

    Pneumothorax after trauma can be a life threatening injury and its care requires expeditious and accurate diagnosis and possible intervention. We performed a prospective, single blinded study with convenience sampling at a Level I trauma center comparing thoracic ultrasound with chest X-ray and CT scan in the detection of traumatic pneumothorax. Trauma patients that received a thoracic ultrasound, chest X-ray, and chest CT scan were included in the study. The chest X-rays were read by a radiologist who was blinded to the thoracic ultrasound results. Then both were compared with CT scan results. One hundred and twenty-five patients had a thoracic ultrasound performed in the 24-month period. Forty-six patients were excluded from the study due to lack of either a chest X-ray or chest CT scan. Of the remaining 79 patients there were 22 positive pneumothorax found by CT and of those 18 (82%) were found on ultrasound and 7 (32%) were found on chest X-ray. The sensitivity of thoracic ultrasound was found to be 81.8 per cent and the specificity was found to be 100 per cent. The sensitivity of chest X-ray was found to be 31.8 per cent and again the specificity was found to be 100 per cent. The negative predictive value of thoracic ultrasound for pneumothorax was 0.934 and the negative predictive value for chest X-ray for pneumothorax was found to be 0.792. We advocate the use of chest ultrasound for detection of pneumothorax in trauma patients.

  16. Patellofemoral joint dysfunction. Combined diagnostic imaging evaluation (X-rays, 3D helical CT and MRI)

    International Nuclear Information System (INIS)

    Carrascosa, P.; Sanchez, F.; Mazzucco, J.; Capanay, C.; Carrascosa, J.

    2000-01-01

    The comprehensive study including 3D helical CT, Magnetic Resonance (MR) and X-ray exams provide a more complete diagnosis than those obtained through the conventional CT. We studied 43 patients with presumptive or certain diagnosis of patellofemoral instability. All the patients were studied by: a) Radiological pair; b) Tomography under extension and flexion, without and with contraction, using a helical CT equipment; and c) MRI STIR sequence in axial plane. The findings were classified as muscular lesion, cartilage lesion, bone and associated lesions (e.g. synovitis), statistically comparing both studies (comprehensive vs. conventional). The results allow us to accept the hypothesis that the comprehensive study provides a more complete diagnosis about the origin of the patellofemoral dysfunction. In 65% of the patients, the conventional study gave a negative result. Only in 35% of the cases the result was positive, but incomplete, showing only 35.3% of the pathological findings detected by the comprehensive study. (author)

  17. X-Ray Microtomography (μCT as a Useful Tool for Visualization and Interpretation of Shear Strength Test Results

    Directory of Open Access Journals (Sweden)

    Stefaniuk Damian

    2015-02-01

    Full Text Available The paper demonstrates the applicability of X-ray microtomography (ìCT to analysis of the results of shear strength examinations of clayey soils. The method of X-ray three-dimensional imaging offers new possibilities in soil testing. The work focuses on a non-destructive method of evaluation of specimen quality used in shear tests and mechanical behavior of soil. The paper presents the results of examination of 4 selected clayey soils. Specimens prepared for the triaxial test have been scanned using ìCT before and after the triaxial compression tests. The shear strength parameters of the soils have been estimated. Changes in soil structure caused by compression and shear failure have been presented as visualizations of the samples tested. This allowed for improved interpretation and evaluation of soil strength parameters and recognition of pre-existing fissures and the exact mode of failure. Basic geometrical parameters have been determined for selected cross-sections of specimens after failure. The test results indicate the utility of the method applied in soil testing.

  18. Physics of x-ray computed tomography

    International Nuclear Information System (INIS)

    Akutagawa, W.M.; Huth, G.C.

    1976-01-01

    Sections are included on theoretical limits of x-ray computed tomography and the relationship of these limits to human organ imaging and specific disease diagnosis; potential of x-ray computed tomography in detection of small calcified particles in early breast cancer detection; early lung cancer measurement and detection; advanced materials for ionizing radiation detection; positron system with circular ring transaxial tomographic camera; contrast mechanism of transmission scanner and algorithms; and status of design on a 200 keV scanning proton microprobe

  19. MiR-34a is up-regulated in response to low dose, low energy X-ray induced DNA damage in breast cells

    International Nuclear Information System (INIS)

    Stankevicins, Luiza; Almeida, Carlos Eduardo de; Moura Gallo, Claudia Vitoria de; Almeida da Silva, Ana Paula; Ventura dos Passos, Flavia; Santos Ferreira, Evelin dos; Menks Ribeiro, Maria Cecilia; G David, Mariano; J Pires, Evandro; Ferreira-Machado, Samara Cristina; Vassetzky, Yegor

    2013-01-01

    MicroRNAs are non-coding RNAs involved in the regulation of gene expression including DNA damage responses. Low doses of low energy X-ray radiation, similar to those used in mammographic exams, has been described to be genotoxic. In the present work we investigated the expression of miR-34a; a well described p53-regulated miRNA implicated in cell responses to X-ray irradiation at low doses. Non-cancerous breast cell line MCF-10A and cancerous T-47D and MCF-7 cell lines were submitted to a low-energy X-ray irradiation (ranging from 28–30 Kv) using a dose of 5 Gy. The expression level of miR-34a, let-7a and miR-21 was assessed by qRT-PCR at 4 and 24 hours post-irradiation. DNA damage was then measured by comet assay and micronuclei estimation in MCF-10A and MCF-7 cell lines, where an increase of miR-34a levels could be observed after irradiation. The rate of apoptotic cells was estimated by nuclear staining and fluorescence microscopy. These experiments were also performed at low doses (3; 12 and 48 mGy) in MCF-10A and MCF-7 cell lines. We have observed an increase in miR-34a expression 4 hours post-irradiation at 5 Gy in MCF-10A and MCF-7 cell lines while its level did not change in T-47D, a breast cancer cell line bearing non-functional p53. At low doses, miR-34a was up-regulated in non-tumoral MCF-10A to a higher extent as compared to MCF-7. MiR-34a levels decreased 24 hours post-irradiation. We have also observed DNA damage and apoptosis at low-energy X-ray irradiation at low doses and the high dose in MCF-10A and MCF-7 4 and 24 hours post-irradiation relative to the mock control. Low energy X-ray is able to promote DNA strand breaks and miR-34a might be involved in cell responses to low energy X-ray DNA damage. MiR-34a expression correlates with X-ray dose, time after irradiation and cell type. The present study reinforces the need of investigating consequences of low dose X-ray irradiation of breast cells

  20. Three dimensional subsurface elemental identification of minerals using confocal micro-X-ray fluorescence and micro-X-ray computed tomography

    International Nuclear Information System (INIS)

    Cordes, Nikolaus L.; Seshadri, Srivatsan; Havrilla, George J.; Yuan, Xiaoli; Feser, Michael; Patterson, Brian M.

    2015-01-01

    Current non-destructive elemental characterization methods, such as scanning electron microscopy-based energy dispersive spectroscopy (SEM–EDS) and micro-X-ray fluorescence spectroscopy (MXRF), are limited to either elemental identification at the surface (SEM–EDS) or suffer from an inability to discriminate between surface or depth information (MXRF). Thus, a non-destructive elemental characterization of individual embedded particles beneath the surface is impossible with either of these techniques. This limitation can be overcome by using laboratory-based 3D confocal micro-X-ray fluorescence spectroscopy (confocal MXRF). This technique utilizes focusing optics on the X-ray source and detector which allows for spatial discrimination in all three dimensions. However, the voxel-by-voxel serial acquisition of a 3D elemental scan can be very time-intensive (~ 1 to 4 weeks) if it is necessary to locate individual embedded particles of interest. As an example, if each point takes a 5 s measurement time, a small volume of 50 × 50 × 50 pixels leads to an acquisition time of approximately 174 h, not including sample stage movement time. Initially screening the samples for particles of interest using micro-X-ray computed tomography (micro-CT) can significantly reduce the time required to spatially locate these particles. Once located, these individual particles can be elementally characterized with confocal MXRF. Herein, we report the elemental identification of high atomic number surface and subsurface particles embedded in a mineralogical matrix by coupling micro-CT and confocal MXRF. Synergistically, these two X-ray based techniques first rapidly locate and then elementally identify individual subsurface particles. - Highlights: • Coupling of confocal X-ray fluorescence spectroscopy and X-ray computed tomography • Qualitative elemental identification of surface and subsurface mineral particles • Non-destructive particle size measurements • Utilization of

  1. Three dimensional subsurface elemental identification of minerals using confocal micro-X-ray fluorescence and micro-X-ray computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Cordes, Nikolaus L., E-mail: ncordes@lanl.gov [Polymers and Coatings Group, Material Science and Technology Division, Los Alamos National Laboratory, Los Alamos, NM 87545 (United States); Seshadri, Srivatsan, E-mail: srivatsan.seshadri@zeiss.com [Carl Zeiss X-ray Microscopy, Inc., Pleasanton, CA 94588 (United States); Havrilla, George J. [Chemical Diagnostics and Engineering, Chemistry Division, Los Alamos National Laboratory, Los Alamos, NM 87545 (United States); Yuan, Xiaoli [Julius Kruttschnitt Mineral Research Centre, University of Queensland, Indooroopilly, Brisbane, QLD 4068 (Australia); Feser, Michael [Carl Zeiss X-ray Microscopy, Inc., Pleasanton, CA 94588 (United States); Patterson, Brian M. [Polymers and Coatings Group, Material Science and Technology Division, Los Alamos National Laboratory, Los Alamos, NM 87545 (United States)

    2015-01-01

    Current non-destructive elemental characterization methods, such as scanning electron microscopy-based energy dispersive spectroscopy (SEM–EDS) and micro-X-ray fluorescence spectroscopy (MXRF), are limited to either elemental identification at the surface (SEM–EDS) or suffer from an inability to discriminate between surface or depth information (MXRF). Thus, a non-destructive elemental characterization of individual embedded particles beneath the surface is impossible with either of these techniques. This limitation can be overcome by using laboratory-based 3D confocal micro-X-ray fluorescence spectroscopy (confocal MXRF). This technique utilizes focusing optics on the X-ray source and detector which allows for spatial discrimination in all three dimensions. However, the voxel-by-voxel serial acquisition of a 3D elemental scan can be very time-intensive (~ 1 to 4 weeks) if it is necessary to locate individual embedded particles of interest. As an example, if each point takes a 5 s measurement time, a small volume of 50 × 50 × 50 pixels leads to an acquisition time of approximately 174 h, not including sample stage movement time. Initially screening the samples for particles of interest using micro-X-ray computed tomography (micro-CT) can significantly reduce the time required to spatially locate these particles. Once located, these individual particles can be elementally characterized with confocal MXRF. Herein, we report the elemental identification of high atomic number surface and subsurface particles embedded in a mineralogical matrix by coupling micro-CT and confocal MXRF. Synergistically, these two X-ray based techniques first rapidly locate and then elementally identify individual subsurface particles. - Highlights: • Coupling of confocal X-ray fluorescence spectroscopy and X-ray computed tomography • Qualitative elemental identification of surface and subsurface mineral particles • Non-destructive particle size measurements • Utilization of

  2. Scatter correction method for x-ray CT using primary modulation: Phantom studies

    International Nuclear Information System (INIS)

    Gao Hewei; Fahrig, Rebecca; Bennett, N. Robert; Sun Mingshan; Star-Lack, Josh; Zhu Lei

    2010-01-01

    Purpose: Scatter correction is a major challenge in x-ray imaging using large area detectors. Recently, the authors proposed a promising scatter correction method for x-ray computed tomography (CT) using primary modulation. Proof of concept was previously illustrated by Monte Carlo simulations and physical experiments on a small phantom with a simple geometry. In this work, the authors provide a quantitative evaluation of the primary modulation technique and demonstrate its performance in applications where scatter correction is more challenging. Methods: The authors first analyze the potential errors of the estimated scatter in the primary modulation method. On two tabletop CT systems, the method is investigated using three phantoms: A Catphan(c)600 phantom, an anthropomorphic chest phantom, and the Catphan(c)600 phantom with two annuli. Two different primary modulators are also designed to show the impact of the modulator parameters on the scatter correction efficiency. The first is an aluminum modulator with a weak modulation and a low modulation frequency, and the second is a copper modulator with a strong modulation and a high modulation frequency. Results: On the Catphan(c)600 phantom in the first study, the method reduces the error of the CT number in the selected regions of interest (ROIs) from 371.4 to 21.9 Hounsfield units (HU); the contrast to noise ratio also increases from 10.9 to 19.2. On the anthropomorphic chest phantom in the second study, which represents a more difficult case due to the high scatter signals and object heterogeneity, the method reduces the error of the CT number from 327 to 19 HU in the selected ROIs and from 31.4% to 5.7% on the overall average. The third study is to investigate the impact of object size on the efficiency of our method. The scatter-to-primary ratio estimation error on the Catphan(c)600 phantom without any annulus (20 cm in diameter) is at the level of 0.04, it rises to 0.07 and 0.1 on the phantom with an

  3. Comparison of x-ray computed tomography, through-transmission ultrasound, and low-kV x-ray imaging for characterizing green-state ceramics

    International Nuclear Information System (INIS)

    Roberts, R.A.; Ellingson, W.A.; Vannier, M.W.

    1985-06-01

    Green-state MgAl 2 O 4 compact disk specimens have been studied by x-ray computed tomography (CT), through-transmission pulsed ultrasound, and low-kV x-ray imaging to compare the abilities of these nondestructive evaluation (NDE) methods to detect flaws and density variations. X-ray computed tomographic images were obtained from a 125-kV (peak) imaging system with a 512 x 512 matrix and a pixel size of 100 μm. A 3- to 10- MHz focused-beam ultrasonic transducer was used, together with special immersion techniques, to obtain topographical maps of acoustic attenuation and phase velocity; a 30 x 30 matrix was used in the ultrasonic scans. A 35-kV x-ray system with high-resolution type RR film was used to obtain conventional radiographs. Large-scale nonuniform density gradients were detected with CT and ultrasonics in supposedly uniform ceramic disks. In addition, inclusions in the green-state samples were detected by all three methods, with each method providing certain advantages. The influence of grain structure and other ceramic powder characteristics will be examined in the future. 5 refs., 9 figs

  4. A Novel Volume CT With X-Ray on a Trough-Like Surface and Point Detectors on Circle-Plus-Arc Curve

    National Research Council Canada - National Science Library

    Xu, H

    2001-01-01

    A novel imaging mode of cone-beam volume CT is proposed in this paper. It adopts a raster scanning x-ray source on a trough-like surface, and a group of point detectors distributing on a large circle plus an orthogonal arc...

  5. WE-DE-207B-10: Library-Based X-Ray Scatter Correction for Dedicated Cone-Beam Breast CT: Clinical Validation

    Energy Technology Data Exchange (ETDEWEB)

    Shi, L; Zhu, L [Georgia Institute of Technology, Atlanta, GA (Georgia); Vedantham, S; Karellas, A [University of Massachusetts Medical School, Worcester, MA (United States)

    2016-06-15

    Purpose: Scatter contamination is detrimental to image quality in dedicated cone-beam breast CT (CBBCT), resulting in cupping artifacts and loss of contrast in reconstructed images. Such effects impede visualization of breast lesions and the quantitative accuracy. Previously, we proposed a library-based software approach to suppress scatter on CBBCT images. In this work, we quantify the efficacy and stability of this approach using datasets from 15 human subjects. Methods: A pre-computed scatter library is generated using Monte Carlo simulations for semi-ellipsoid breast models and homogeneous fibroglandular/adipose tissue mixture encompassing the range reported in literature. Projection datasets from 15 human subjects that cover 95 percentile of breast dimensions and fibroglandular volume fraction were included in the analysis. Our investigations indicate that it is sufficient to consider the breast dimensions alone and variation in fibroglandular fraction does not significantly affect the scatter-to-primary ratio. The breast diameter is measured from a first-pass reconstruction; the appropriate scatter distribution is selected from the library; and, deformed by considering the discrepancy in total projection intensity between the clinical dataset and the simulated semi-ellipsoidal breast. The deformed scatter-distribution is subtracted from the measured projections for scatter correction. Spatial non-uniformity (SNU) and contrast-to-noise ratio (CNR) were used as quantitative metrics to evaluate the results. Results: On the 15 patient cases, our method reduced the overall image spatial non-uniformity (SNU) from 7.14%±2.94% (mean ± standard deviation) to 2.47%±0.68% in coronal view and from 10.14%±4.1% to 3.02% ±1.26% in sagittal view. The average contrast to noise ratio (CNR) improved by a factor of 1.49±0.40 in coronal view and by 2.12±1.54 in sagittal view. Conclusion: We demonstrate the robustness and effectiveness of a library-based scatter correction

  6. TU-EF-207-00: Advances in Breast Imaging

    Energy Technology Data Exchange (ETDEWEB)

    NONE

    2015-06-15

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation.

  7. A case of recurrence-mimicking charcoal granuloma in a breast cancer patient: Ultrasound,CT, PET/CT and breast-specific gamma imaging findings

    Energy Technology Data Exchange (ETDEWEB)

    Park, Dae Woong; Park, Ji Yeon; Park, Noh Hyuck; Kim, Seon Jeong; Shin, Hyuck Jai; Lee, Jeong Ju [Myongji Hospital, Seonam University College of Medicine, Goyang (Korea, Republic of); Yi, Seong Yoon [Div. of Hematology-Oncology, Dept. of Internal Medicine, Inje University Ilsan Paik Hospital, Goyang (Korea, Republic of)

    2016-07-15

    Charcoal remains stable without causing a foreign body reaction and it may be used for preoperative localization of a non-palpable breast mass. However, cases of post-charcoal-marking granuloma have only rarely been reported in the breast, and a charcoal granuloma can be misdiagnosed as malignancy. Herein, we report the ultrasound, computed tomography (CT), 18F-fluorodeoxyglucose-positron emission tomography/CT, and breast-specific gamma imaging findings of recurrence-mimicking charcoal granuloma after breast conserving surgery, following localization with charcoal in a breast cancer patient.

  8. Quantification and normalization of x-ray mammograms

    International Nuclear Information System (INIS)

    Tromans, Christopher E; Cocker, Mary R; Brady, Michael

    2012-01-01

    The analysis of (x-ray) mammograms remains qualitative, relying on the judgement of clinicians. We present a novel method to compute a quantitative, normalized measure of tissue radiodensity traversed by the primary beam incident on each pixel of a mammogram, a measure we term the standard attenuation rate (SAR). SAR enables: the estimation of breast density which is linked to cancer risk; direct comparison between images; the full potential of computer aided diagnosis to be utilized; and a basis for digital breast tomosynthesis reconstruction. It does this by removing the effects of the imaging conditions under which the mammogram is acquired. First, the x-ray spectrum incident upon the breast is calculated, and from this, the energy exiting the breast is calculated. The contribution of scattered radiation is calculated and subtracted. The SAR measure is the scaling factor that must be applied to the reference material in order to match the primary attenuation of the breast. Specifically, this is the scaled reference material attenuation which when traversed by an identical beam to that traversing the breast, and when subsequently detected, results in the primary component of the pixel intensity observed in the breast image. We present results using two tissue equivalent phantoms, as well as a sensitivity analysis to detector response changes over time and possible errors in compressed thickness measurement. (paper)

  9. X-ray evaluation of bone tissue density in rheumatoid arthritis

    International Nuclear Information System (INIS)

    Kotlubej, O.V.; Kamenets'kij, M.S.; Tkachenko, G.D.; Momot, N.V.

    1998-01-01

    Densitometry of hand and foot films of 111 patients with rheumatoid arthritis was performed. Twelve patients underwent CT of lumbar vertebra with tomodensitometry followed by analysis of the obtained graphic histograms. Photo densitometry of plain x-ray films and histographic analysis of CT scans allows to perform quantitative evaluation of the bone tissue density in rheumatoid arthritis and to reveal osteoporosis at early stages when x-ray features of the disease are absent

  10. Theoretical estimation of proton induced X-ray emission yield of the trace elements present in the lung and breast cancer

    International Nuclear Information System (INIS)

    Manjunatha, H.C.; Sowmya, N.

    2013-01-01

    X-rays may be produced following the excitation of target atoms induced by an energetic incident ion beam of protons. Proton induced X-ray emission (PIXE) analysis has been used for many years for the determination of elemental composition of materials using X-rays. Recent interest in the proton induced X-ray emission cross section has arisen due to their importance in the rapidly expanding field of PIXE analysis. One of the steps in the analysis is to fit the measured X-ray spectrum with theoretical spectrum. The theoretical cross section and yields are essential for the evaluation of spectrum. We have theoretically evaluated the PIXE cross sections for trace elements in the lung and breast cancer tissues such as Cl, K, Ca,Ti, Cr, Mn, Fe, Ni, Cu, Zn, As, Se, Br, Rb, P, S, Sr, Hg and Pb. The estimated cross section is used in the evaluation of Proton induced X-ray emission spectrum for the given trace elements.We have also evaluated the Proton induced X-ray emission yields in the thin and thick target of the given trace elements. The evaluated Proton induced X-ray emission cross-section, spectrum and yields are graphically represented. Some of these values are also tabulated. Proton induced X-ray emission cross sections and a yield for the given trace elements varies with the energy. PIXE yield depends on a real density and does not on thickness of the target. (author)

  11. Effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies

    International Nuclear Information System (INIS)

    Murase, Kenya; Nanjo, Takafumi; Ii, Satoshi; Miyazaki, Shohei; Hirata, Masaaki; Sugawara, Yoshifumi; Kudo, Masayuki; Sasaki, Kousuke; Mochizuki, Teruhito

    2005-01-01

    The purpose of this study was to investigate the effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies using multi-detector row CT (MDCT). Following the standard CT perfusion study protocol, continuous (cine) scans (1 s/rotation x 60 s) consisting of four 5 mm thick contiguous slices were performed using an MDCT scanner with a tube voltage of 80 kVp and a tube current of 200 mA. We generated the simulated images with tube currents of 50 mA, 100 mA and 150 mA by adding the corresponding noise to the raw scan data of the original image acquired above using a noise simulation tool. From the original and simulated images, we generated the functional images of cerebral blood flow (CBF), cerebral blood volume (CBV) and mean transit time (MTT) in seven patients with cerebrovascular disease, and compared the correlation coefficients (CCs) between the perfusion parameter values obtained from the original and simulated images. The coefficients of variation (CVs) in the white matter were also compared. The CC values deteriorated with decreasing tube current. There was a significant difference between 50 mA and 100 mA for all perfusion parameters. The CV values increased with decreasing tube current. There were significant differences between 50 mA and 100 mA and between 100 mA and 150 mA for CBF. For CBV and MTT, there was also a significant difference between 150 mA and 200 mA. This study will be useful for understanding the effect of x-ray tube current on the accuracy of cerebral perfusion parameters obtained by CT perfusion studies using MDCT, and for selecting the tube current

  12. X-ray imaging with monochromatic synchrotron radiation. Fluorescent and phase-contrast method

    Energy Technology Data Exchange (ETDEWEB)

    Takeda, Tohoru; Itai, Yuji [Tsukuba Univ., Ibaraki (Japan). Inst. of Clinical Medicine

    2002-05-01

    To obtain the high sensitive x-ray images of biomedical object, new x-ray imaging techniques using fluorescent x-ray and phase-contrast x-ray are being developed in Japan. Fluorescent x-ray CT can detect very small amounts of specific elements in the order of ppm at one pixel, whereas phase-contrast x-ray imaging with interferometer can detect minute differences of biological object. Here, our recent experimental results are presented. (author)

  13. Scintillating ribbon x-ray detector

    International Nuclear Information System (INIS)

    Kinchen, B.E.; Rogers, A.

    1995-01-01

    A patent in the early 1970's by Aerojet Corporation in Sacramento, CA put forth the idea of using an array of scintillating fibers for x-ray detection and imaging. In about 1975, Pratt and Whitney Aircraft in East Hartford, CT designed and manufactured an imaging system based on the patent. The device was 1.75 in thick in the direction of the x-ray beam and about 4 in. by 4 in. square. The device was used with a 8 MeV x-ray source to image and measure internal clearances within operating aircraft, gas turbines engines. There are significant advantages of fiber optic detectors in x-ray detection. However, the advantages are often outweighed by the disadvantages. Two of the advantages of scintillating fiber optic x-ray detectors are: (1) high limiting spatial frequency -- between 20 and 25 lp/mm; and (2) excellent x-ray stopping power -- they can be made thick and retain spatial resolution. In traditional fiber optic detectors the x-rays are oriented parallel to the long axis of the fiber. For the scintillating ribbon x-ray sensor, the x-rays are oriented normal to the fiber long axis. This ribbon sensor technique has a number of advantages over the two current radiographic techniques digital x-radiography and x-ray film: The main advantage the ribbon has is size and shape. It can be as thin as 0.05 in., virtually any width or length, and flexible. Once positioned in a given location, 20 to 100 square inches of the object being inspected can be imaged with a single x-ray beam sweep. It is clear that conventional digital cameras do not lend themselves to placement between walls of aircraft structures or similar items requiring x-ray inspections. A prototype scintillating ribbon x-ray sensor has been fabricated and tested by Synergistic Detector Designs. Images were acquired on corrosion test panels of aluminum fabricated by Iowa State University

  14. The study on the X-ray correction method of long fracture displacement

    International Nuclear Information System (INIS)

    Jia Bin; Huang Ailing; Chen Fuzhong; Men Chunyan; Sui Chengzong; Cui Yiming; Yang Yundong

    2010-01-01

    Objective: To explore the image correction of fracture displacement by conventional X-ray photography (ortho tropic and lateral) and test by computed tomography (CT). Methods: The correction method of fracture displacement was designed according to geometry of X-ray photography. Selected one midhumeral fracture specimen which designed with lateral shift and angular displacement, and scanned from anteroposterior and position respectively, and also volume scanned using CT, the data obtained from volume scan were processed using multiplanar reconstruction (MPR) and shaded surface display (SSD). The displacement data relied on X-ray image, CT with MPR and SSD processing, actual design of specimens were compared respectively. Results: The direction and degree of displacement among correction data of X-ray images and the data from MPR and SSD, actual design of specimen were little difference, location difference <1.5 mm, degree difference <1.5 degree. Conclusion: It is really reliable for fracture displacement by conventional X-ray photography with coordinate correction, and it is helpful to obviously improve the diagnostic accuracy of the degree of fracture displacement. (authors)

  15. Analysis of iterative region-of-interest image reconstruction for x-ray computed tomography.

    Science.gov (United States)

    Sidky, Emil Y; Kraemer, David N; Roth, Erin G; Ullberg, Christer; Reiser, Ingrid S; Pan, Xiaochuan

    2014-10-03

    One of the challenges for iterative image reconstruction (IIR) is that such algorithms solve an imaging model implicitly, requiring a complete representation of the scanned subject within the viewing domain of the scanner. This requirement can place a prohibitively high computational burden for IIR applied to x-ray computed tomography (CT), especially when high-resolution tomographic volumes are required. In this work, we aim to develop an IIR algorithm for direct region-of-interest (ROI) image reconstruction. The proposed class of IIR algorithms is based on an optimization problem that incorporates a data fidelity term, which compares a derivative of the estimated data with the available projection data. In order to characterize this optimization problem, we apply it to computer-simulated two-dimensional fan-beam CT data, using both ideal noiseless data and realistic data containing a level of noise comparable to that of the breast CT application. The proposed method is demonstrated for both complete field-of-view and ROI imaging. To demonstrate the potential utility of the proposed ROI imaging method, it is applied to actual CT scanner data.

  16. Development of a fluorescent x-ray source for medical imaging

    Science.gov (United States)

    Toyofuku, F.; Tokumori, K.; Nishimura, K.; Saito, T.; Takeda, T.; Itai, Y.; Hyodo, K.; Ando, M.; Endo, M.; Naito, H.; Uyama, C.

    1995-02-01

    A fluorescent x-ray source for medical imaging, such as K-edge subtraction angiography and monochromatic x-ray CT, has been developed. Using a 6.5 GeV accumulation ring in Tsukuba, fluorescent x rays, which range from about 30 to 70 keV are generated by irradiating several target materials. Measurements have been made of output intensities and energy spectra for different target angles and extraction angles. The intensities of fluorescent x rays at a 30 mA beam current are on the order of 1-3×106 photons/mm2/s at 30 cm from the local spot where the incident beam is collimated to 1 mm2. A phantom which contains three different contrast media (iodine, barium, gadolinium) was used for the K-edge energy subtraction, and element selective CT images were obtained.

  17. X-ray Computed Tomography Image Quality Indicator (IQI) Development

    Data.gov (United States)

    National Aeronautics and Space Administration — Phase one of the program is to identify suitable x-ray Computed Tomography (CT) Image Quality Indicator (IQI) design(s) that can be used to adequately capture CT...

  18. Photon Counting Energy Dispersive Detector Arrays for X-ray Imaging.

    Science.gov (United States)

    Iwanczyk, Jan S; Nygård, Einar; Meirav, Oded; Arenson, Jerry; Barber, William C; Hartsough, Neal E; Malakhov, Nail; Wessel, Jan C

    2009-01-01

    The development of an innovative detector technology for photon-counting in X-ray imaging is reported. This new generation of detectors, based on pixellated cadmium telluride (CdTe) and cadmium zinc telluride (CZT) detector arrays electrically connected to application specific integrated circuits (ASICs) for readout, will produce fast and highly efficient photon-counting and energy-dispersive X-ray imaging. There are a number of applications that can greatly benefit from these novel imagers including mammography, planar radiography, and computed tomography (CT). Systems based on this new detector technology can provide compositional analysis of tissue through spectroscopic X-ray imaging, significantly improve overall image quality, and may significantly reduce X-ray dose to the patient. A very high X-ray flux is utilized in many of these applications. For example, CT scanners can produce ~100 Mphotons/mm(2)/s in the unattenuated beam. High flux is required in order to collect sufficient photon statistics in the measurement of the transmitted flux (attenuated beam) during the very short time frame of a CT scan. This high count rate combined with a need for high detection efficiency requires the development of detector structures that can provide a response signal much faster than the transit time of carriers over the whole detector thickness. We have developed CdTe and CZT detector array structures which are 3 mm thick with 16×16 pixels and a 1 mm pixel pitch. These structures, in the two different implementations presented here, utilize either a small pixel effect or a drift phenomenon. An energy resolution of 4.75% at 122 keV has been obtained with a 30 ns peaking time using discrete electronics and a (57)Co source. An output rate of 6×10(6) counts per second per individual pixel has been obtained with our ASIC readout electronics and a clinical CT X-ray tube. Additionally, the first clinical CT images, taken with several of our prototype photon-counting and

  19. Optimized control of X-ray exposure and image noise using a particular multislice CT scanner

    International Nuclear Information System (INIS)

    Yamamoto, Shuji; Suzuki, Masahiro; Kakinuma, Ryutaro; Moriyama, Noriyuki; Koyama, Yoshihiro; Nagasawa, Hirofumi

    2008-01-01

    Patient dose reduction in computed tomography (CT) always results in a trade off between radiation exposure and image quality. There are few reports that estimate the relationship between image quality and X-ray exposure in CT examinations as one optimal index. The purpose of this study was to determine the optimal parameter settings enabling a low radiation exposure without compromising image quality using a particular 4-row multislice CT (MSCT) scanner (Aquilion VZ 4-slice CT scanner, Toshiba Medical Systems Corporation, Otawara, Tochigi, Japan). Normalized dose divided by image noise for helical pitches (nDNR: normalized dose to noise ratio) were calculated in consideration of beam collimation and tube current-time product. Optimal tube current-time product was calculated using the nDNR for the helical pitches based on user-defined standards of quality of the CT image. As a result, the nDNR proved to be well-supported to decrease the patient exposure in various exposure conditions of MSCT scans; however, the dose and image noise did not show a linear relation to the helical pitch. In conclusion, nDNR can be applied to patient dose reduction while keeping an acceptable image quality using a particular 4-row MSCT scanner. (author)

  20. Low-dose CT of the paranasal sinuses. Minimizing X-ray exposure with spectral shaping

    Energy Technology Data Exchange (ETDEWEB)

    Wuest, Wolfgang [Friedrich-Alexander-University Erlangen-Nuremberg, Radiological Institute, Erlangen (Germany); Radiological Institute, Erlangen (Germany); May, Matthias; Saake, Marc; Brand, Michael; Uder, Michael; Lell, Michael [Friedrich-Alexander-University Erlangen-Nuremberg, Radiological Institute, Erlangen (Germany)

    2016-11-15

    Shaping the energy spectrum of the X-ray beam has been shown to be beneficial in low-dose CT. This study's aim was to investigate dose and image quality of tin filtration at 100 kV for pre-operative planning in low-dose paranasal CT imaging in a large patient cohort. In a prospective trial, 129 patients were included. 64 patients were randomly assigned to the study protocol (100 kV with additional tin filtration, 150mAs, 192 x 0.6-mm slice collimation) and 65 patients to the standard low-dose protocol (100 kV, 50mAs, 128 x 0.6-mm slice collimation). To assess the image quality, subjective parameters were evaluated using a five-point scale. This scale was applied on overall image quality and contour delineation of critical anatomical structures. All scans were of diagnostic image quality. Bony structures were of good diagnostic image quality in both groups, soft tissues were of sufficient diagnostic image quality in the study group because of a high level of noise. Radiation exposure was very low in both groups, but significantly lower in the study group (CTDI{sub vol} 1.2 mGy vs. 4.4 mGy, p < 0.001). Spectral optimization (tin filtration at 100 kV) allows for visualization of the paranasal sinus with sufficient image quality at a very low radiation exposure. (orig.)

  1. Amorphous In-Ga-Zn-O thin-film transistor active pixel sensor x-ray imager for digital breast tomosynthesis.

    Science.gov (United States)

    Zhao, Chumin; Kanicki, Jerzy

    2014-09-01

    The breast cancer detection rate for digital breast tomosynthesis (DBT) is limited by the x-ray image quality. The limiting Nyquist frequency for current DBT systems is around 5 lp/mm, while the fine image details contained in the high spatial frequency region (>5 lp/mm) are lost. Also today the tomosynthesis patient dose is high (0.67-3.52 mGy). To address current issues, in this paper, for the first time, a high-resolution low-dose organic photodetector/amorphous In-Ga-Zn-O thin-film transistor (a-IGZO TFT) active pixel sensor (APS) x-ray imager is proposed for next generation DBT systems. The indirect x-ray detector is based on a combination of a novel low-cost organic photodiode (OPD) and a cesium iodide-based (CsI:Tl) scintillator. The proposed APS x-ray imager overcomes the difficulty of weak signal detection, when small pixel size and low exposure conditions are used, by an on-pixel signal amplification with a significant charge gain. The electrical performance of a-IGZO TFT APS pixel circuit is investigated by SPICE simulation using modified Rensselaer Polytechnic Institute amorphous silicon (a-Si:H) TFT model. Finally, the noise, detective quantum efficiency (DQE), and resolvability of the complete system are modeled using the cascaded system formalism. The result demonstrates that a large charge gain of 31-122 is achieved for the proposed high-mobility (5-20 cm2/V s) amorphous metal-oxide TFT APS. The charge gain is sufficient to eliminate the TFT thermal noise, flicker noise as well as the external readout circuit noise. Moreover, the low TFT (sensor imager under 1 mR, indicating good image quality under low dose. A threefold reduction of current tomosynthesis dose is expected if proposed technology is combined with an advanced DBT image reconstruction method. The proposed a-IGZO APS x-ray imager with a pixel pitch6.67 lp/mm) and a low dose (<0.4 mGy) in next generation DBT systems.

  2. Imaging-therapy computed tomography with quasi-monochromatic X-rays.

    Science.gov (United States)

    Jost, Gregor; Golfier, Sven; Lawaczeck, Ruediger; Weinmann, Hanns-Joachim; Gerlach, Martin; Cibik, Levent; Krumrey, Michael; Fratzscher, Daniel; Rabe, Johannis; Arkadiev, Vladimir; Haschke, Michael; Langhoff, Norbert; Wedell, Reiner; Luedemann, Lutz; Wust, Peter; Pietsch, Hubertus

    2008-12-01

    Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer. The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified. The X-ray optical module filters selectively the energy of the tungsten K alpha-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine. An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT.

  3. A software-based x-ray scatter correction method for breast tomosynthesis

    International Nuclear Information System (INIS)

    Jia Feng, Steve Si; Sechopoulos, Ioannis

    2011-01-01

    Purpose: To develop a software-based scatter correction method for digital breast tomosynthesis (DBT) imaging and investigate its impact on the image quality of tomosynthesis reconstructions of both phantoms and patients. Methods: A Monte Carlo (MC) simulation of x-ray scatter, with geometry matching that of the cranio-caudal (CC) view of a DBT clinical prototype, was developed using the Geant4 toolkit and used to generate maps of the scatter-to-primary ratio (SPR) of a number of homogeneous standard-shaped breasts of varying sizes. Dimension-matched SPR maps were then deformed and registered to DBT acquisition projections, allowing for the estimation of the primary x-ray signal acquired by the imaging system. Noise filtering of the estimated projections was then performed to reduce the impact of the quantum noise of the x-ray scatter. Three dimensional (3D) reconstruction was then performed using the maximum likelihood-expectation maximization (MLEM) method. This process was tested on acquisitions of a heterogeneous 50/50 adipose/glandular tomosynthesis phantom with embedded masses, fibers, and microcalcifications and on acquisitions of patients. The image quality of the reconstructions of the scatter-corrected and uncorrected projections was analyzed by studying the signal-difference-to-noise ratio (SDNR), the integral of the signal in each mass lesion (integrated mass signal, IMS), and the modulation transfer function (MTF). Results: The reconstructions of the scatter-corrected projections demonstrated superior image quality. The SDNR of masses embedded in a 5 cm thick tomosynthesis phantom improved 60%-66%, while the SDNR of the smallest mass in an 8 cm thick phantom improved by 59% (p < 0.01). The IMS of the masses in the 5 cm thick phantom also improved by 15%-29%, while the IMS of the masses in the 8 cm thick phantom improved by 26%-62% (p < 0.01). Some embedded microcalcifications in the tomosynthesis phantoms were visible only in the scatter

  4. Evaluation of the Significance of Incidental Breast Lesions Detected by Chest CT

    International Nuclear Information System (INIS)

    Kim, Jae Hyun; Chang, Yun Woo; Hwang, Jung Hwa; Kim, Hyung Hwan; Lee, Eun Hye; Yang, Seung Boo

    2013-01-01

    To evaluate the significance of incidentally detected breast lesions on a chest CT scan. Thirty-six incidental breast lesions in 26 patients were detected on a chest CT scan and were correlated with breast sonography, retrospectively. Among them, twenty-four breast lesions in 20 patients that were correlated with chest CT and sonography were available to pathology or follow up sonography. The CT findings were compared with sonographic findings according to the pathologic results. Incidentally detected breast lesions on a chest CT scan were correlated with sonography in 86% (31/36). Among 24 lesions that were available to pathology or follow up sonography, seven (29.2%) lesions were malignant and 17 (70.8%) lesions were benign. CT revealed a significant difference between benign and malignant lesions in terms of shape and margin (p = 0.007; p = 0.008, respectively). The CT findings were well correlated with sonographic findings in shape and margin (p = 0.001, respectively). Incidentally detected breast lesions on chest CT can be correlated with sonography. An irregular shape or a non-circumscribed margin of breast lesions on a CT scan can be considered as a suggestive sign of malignancy.

  5. X-ray strain tensor imaging: FEM simulation and experiments with a micro-CT.

    Science.gov (United States)

    Kim, Jae G; Park, So E; Lee, Soo Y

    2014-01-01

    In tissue elasticity imaging, measuring the strain tensor components is necessary to solve the inverse problem. However, it is impractical to measure all the tensor components in ultrasound or MRI elastography because of their anisotropic spatial resolution. The objective of this study is to compute 3D strain tensor maps from the 3D CT images of a tissue-mimicking phantom. We took 3D micro-CT images of the phantom twice with applying two different mechanical compressions to it. Applying the 3D image correlation technique to the CT images under different compression, we computed 3D displacement vectors and strain tensors at every pixel. To evaluate the accuracy of the strain tensor maps, we made a 3D FEM model of the phantom, and we computed strain tensor maps through FEM simulation. Experimentally obtained strain tensor maps showed similar patterns to the FEM-simulated ones in visual inspection. The correlation between the strain tensor maps obtained from the experiment and the FEM simulation ranges from 0.03 to 0.93. Even though the strain tensor maps suffer from high level noise, we expect the x-ray strain tensor imaging may find some biomedical applications such as malignant tissue characterization and stress analysis inside the tissues.

  6. Hafnium-Based Contrast Agents for X-ray Computed Tomography.

    Science.gov (United States)

    Berger, Markus; Bauser, Marcus; Frenzel, Thomas; Hilger, Christoph Stephan; Jost, Gregor; Lauria, Silvia; Morgenstern, Bernd; Neis, Christian; Pietsch, Hubertus; Sülzle, Detlev; Hegetschweiler, Kaspar

    2017-05-15

    Heavy-metal-based contrast agents (CAs) offer enhanced X-ray absorption for X-ray computed tomography (CT) compared to the currently used iodinated CAs. We report the discovery of new lanthanide and hafnium azainositol complexes and their optimization with respect to high water solubility and stability. Our efforts culminated in the synthesis of BAY-576, an uncharged hafnium complex with 3:2 stoichiometry and broken complex symmetry. The superior properties of this asymmetrically substituted hafnium CA were demonstrated by a CT angiography study in rabbits that revealed excellent signal contrast enhancement.

  7. Study on data acquisition circuit used in SSPA linear array detector X-ray detection

    CERN Document Server

    Wei Biao; Che Zhen Ping

    2002-01-01

    After SSPA used as X-ray array detector is developed, the authors take a research on the data acquisition circuit applied to the detector. The experiment designed has verified the feasibility of application of this array detector and its data acquisition circuit to X-ray computed tomography (X-CT). The preliminary test results indicate that the method of the X-ray detection is feasible for industry X-CT nondestructive testing, which brings about advantage for detecting and measuring with high resolution, good efficiency and low cost

  8. The significance of X-ray mammography for breast cancer screening

    International Nuclear Information System (INIS)

    Schulz-Wendtland, R.; Kraemer, S.; Doeinghaus, K.; Saebel, M.; Lang, N.; Bautz, W.

    1997-01-01

    One high-priority health problem in our society are the high, age-specifically standardized mortality rates of breast cancer. Breast-conserving therapies are increasingly applied today whenever appropriate, as the patient outcome depends less on the local size of the tumor than on the existing micrometastases and general spreading of the tumor in terms of a systemic tumor lesion at the time of diagnosis and primary therapy. Among the available diagnostic methods, X-ray mammography is the modality of choice for detection of small, pre-clinical carcinoma or pre-invasive stages. Although one should not overestimate the true correct findings in detection of tumors less than 1 cm of size, the proportion of correct findings indicating T1-tumors or minimal cancer obtained in multiple screening projects is surprising. The fact that the number of detected cases involving the lymph nodes is 50% in patients with clinical symptoms and only 20% or even less in diognoses obtained by screening gives reason enough to speak of ''early detection'' in the latter case. Finally, the results of the HIP study, (reduction of mortality rates for women over 50 by 40%), and of the Oestergoetland study (for women between 40 and 70 years of age, 31%) are very convincing. Furthermore, a meta-analysis on the value of breast cancer screening reveals a mortality reduction of 22% in women between 40 and 49 years of age. Further improvement of the mammographic method and quality control in medical evaluation hopefully will contribute to more widespread screening activities also in Germany, and hence to a further reduction of mortality rates. (Orig./vhe) [de

  9. Improvement of analysis precision upon the atomic number and electron density measurement by the dual x-ray CT

    Science.gov (United States)

    Imura, Yukino; Morii, Hisashi; Koike, Akifumi; Okunoyama, Takaharu; Neo, Yoichiro; Mimura, Hidenori; Aoki, Toru

    2010-08-01

    To identify the factor impairing the material identification parameters, which is provided by the dual-energy X-ray computed tomography method using a conventional X-ray tube and a CdTe detector, linear attenuation coefficient was measured by the radioactivity of radio isotopes and compared with theoretical figure. In our study, the atomic number and the electron density is calculated from the linear attenuation coefficient obtained in CT measurement by 64-channel CdTe line detector. To estimate accuracy of CdTe line sensor, it is needed to obtain the linear attenuation coefficient accurately. Using a single detector, the linear attenuation coefficient is verified for accuracy. The energy resolution of CdTe detectors and the method of reconstruction are discussed.

  10. A hyperspectral X-ray computed tomography system for enhanced material identification

    Science.gov (United States)

    Wu, Xiaomei; Wang, Qian; Ma, Jinlei; Zhang, Wei; Li, Po; Fang, Zheng

    2017-08-01

    X-ray computed tomography (CT) can distinguish different materials according to their absorption characteristics. The hyperspectral X-ray CT (HXCT) system proposed in the present work reconstructs each voxel according to its X-ray absorption spectral characteristics. In contrast to a dual-energy or multi-energy CT system, HXCT employs cadmium telluride (CdTe) as the x-ray detector, which provides higher spectral resolution and separate spectral lines according to the material's photon-counter working principle. In this paper, a specimen containing ten different polymer materials randomly arranged was adopted for material identification by HXCT. The filtered back-projection algorithm was applied for image and spectral reconstruction. The first step was to sort the individual material components of the specimen according to their cross-sectional image intensity. The second step was to classify materials with similar intensities according to their reconstructed spectral characteristics. The results demonstrated the feasibility of the proposed material identification process and indicated that the proposed HXCT system has good prospects for a wide range of biomedical and industrial nondestructive testing applications.

  11. Temporal dynamics for soil aggregates determined using X-ray CT scanning

    DEFF Research Database (Denmark)

    Garbout, Amin; Munkholm, Lars Juhl; Hansen, Søren Baarsgaard

    2013-01-01

    Soil structure plays a key role in the ability of soil to fulfil essential soil functions and services in relation to e.g. root growth, gas and water transport and organic matter turnover. However, soils are not a very easy object to study as they are highly complex and opaque to the human eye...... aggregate properties such as volume, surface area and sphericity based on 3D images. We tested the methods on aggregates from different treatments and quantified changes over time. A total of 32 collections of aggregates, enclosed in mesocosms, were incubated in soil to follow the structural changes over....... Traditionally, they have been studied using invasive or destructive techniques. The advantage of using X-ray computed tomography (CT) in soil morphology is that it enables non-destructive quantification of soil structure in three dimensions (3D). The prime objective of the present study was to characterize soil...

  12. Application of micron X-ray CT based on micro-PIXE to investigate the distribution of Cs in silt particles for environmental remediation in Fukushima Prefecture

    Energy Technology Data Exchange (ETDEWEB)

    Ishii, Keizo, E-mail: keizo.ishii@qse.tohoku.ac.jp [Research Center for Remediation Engineering of Environments Contaminated with Radioisotopes, Graduate School of Engineering, Tohoku University, 6-6-01-2, Aramaki Aza-Aoba, Aoba-ku, Sendai 980-8579 (Japan); Hatakeyama, Taisuke; Itoh, Shin; Sata, Daichi [Department of Quantum Science and Energy Engineering, Tohoku University, 6-6-01-2, Aramaki Aza-Aoba, Aoba-ku, Sendai 980-8579 (Japan); Ohnuma, Tohru; Yamaguchi, Toshiro; Arai, Hiromu; Arai, Hirotsugu [Research Center for Remediation Engineering of Environments Contaminated with Radioisotopes, Graduate School of Engineering, Tohoku University, 6-6-01-2, Aramaki Aza-Aoba, Aoba-ku, Sendai 980-8579 (Japan); Matsuyama, Shigeo; Terakawa, Atsuki; Kim, Seong-Yun [Department of Quantum Science and Energy Engineering, Tohoku University, 6-6-01-2, Aramaki Aza-Aoba, Aoba-ku, Sendai 980-8579 (Japan)

    2016-03-15

    We used X-ray computed tomography (CT) using characteristic X-rays produced in micro-particle-induced X-ray emission (PIXE) to investigate the internal structure of silt particles and develop new methods to decontaminate soil containing radioactive cesium. We obtained 3D attenuation coefficient images of silt particles with a diameter of approximately 100 μm for V K and Cr K X-rays. Owing to the absorption edges of the Cs L-shell, the differences between the V K and Cr K X-ray images revealed the spatial distribution of Cs atoms in the silt particles. Cs atoms were distributed over the surfaces of the silt particles to a thickness of approximately 10 μm. This information is useful for the decontamination of silt contaminated by radiation from the Fukushima Daiichi nuclear disaster.

  13. Higher-order scene statistics of breast images

    Science.gov (United States)

    Abbey, Craig K.; Sohl-Dickstein, Jascha N.; Olshausen, Bruno A.; Eckstein, Miguel P.; Boone, John M.

    2009-02-01

    Researchers studying human and computer vision have found description and construction of these systems greatly aided by analysis of the statistical properties of naturally occurring scenes. More specifically, it has been found that receptive fields with directional selectivity and bandwidth properties similar to mammalian visual systems are more closely matched to the statistics of natural scenes. It is argued that this allows for sparse representation of the independent components of natural images [Olshausen and Field, Nature, 1996]. These theories have important implications for medical image perception. For example, will a system that is designed to represent the independent components of natural scenes, where objects occlude one another and illumination is typically reflected, be appropriate for X-ray imaging, where features superimpose on one another and illumination is transmissive? In this research we begin to examine these issues by evaluating higher-order statistical properties of breast images from X-ray projection mammography (PM) and dedicated breast computed tomography (bCT). We evaluate kurtosis in responses of octave bandwidth Gabor filters applied to PM and to coronal slices of bCT scans. We find that kurtosis in PM rises and quickly saturates for filter center frequencies with an average value above 0.95. By contrast, kurtosis in bCT peaks near 0.20 cyc/mm with kurtosis of approximately 2. Our findings suggest that the human visual system may be tuned to represent breast tissue more effectively in bCT over a specific range of spatial frequencies.

  14. Breast tumor segmentation in high resolution x-ray phase contrast analyzer based computed tomography.

    Science.gov (United States)

    Brun, E; Grandl, S; Sztrókay-Gaul, A; Barbone, G; Mittone, A; Gasilov, S; Bravin, A; Coan, P

    2014-11-01

    Phase contrast computed tomography has emerged as an imaging method, which is able to outperform present day clinical mammography in breast tumor visualization while maintaining an equivalent average dose. To this day, no segmentation technique takes into account the specificity of the phase contrast signal. In this study, the authors propose a new mathematical framework for human-guided breast tumor segmentation. This method has been applied to high-resolution images of excised human organs, each of several gigabytes. The authors present a segmentation procedure based on the viscous watershed transform and demonstrate the efficacy of this method on analyzer based phase contrast images. The segmentation of tumors inside two full human breasts is then shown as an example of this procedure's possible applications. A correct and precise identification of the tumor boundaries was obtained and confirmed by manual contouring performed independently by four experienced radiologists. The authors demonstrate that applying the watershed viscous transform allows them to perform the segmentation of tumors in high-resolution x-ray analyzer based phase contrast breast computed tomography images. Combining the additional information provided by the segmentation procedure with the already high definition of morphological details and tissue boundaries offered by phase contrast imaging techniques, will represent a valuable multistep procedure to be used in future medical diagnostic applications.

  15. Imaging-therapy computed tomography with quasi-monochromatic X-rays

    International Nuclear Information System (INIS)

    Jost, Gregor; Golfier, Sven; Lawaczeck, Ruediger; Weinmann, Hanns-Joachim; Gerlach, Martin; Cibik, Levent; Krumrey, Michael; Fratzscher, Daniel; Rabe, Johannis; Arkadiev, Vladimir; Haschke, Michael; Langhoff, Norbert; Wedell, Reiner

    2008-01-01

    Introduction: Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer. Materials and methods: The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified. Results: The X-ray optical module filters selectively the energy of the tungsten Kα-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine. Conclusion: An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT

  16. Imaging-therapy computed tomography with quasi-monochromatic X-rays

    Energy Technology Data Exchange (ETDEWEB)

    Jost, Gregor [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: gregor.jost@bayerhealthcare.com; Golfier, Sven [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: sven.golfier@bayerhealthcare.com; Lawaczeck, Ruediger [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: ruediger.lawaczeck@bayerhealthcare.com; Weinmann, Hanns-Joachim [Bayer Schering Pharma AG, Contrast Media Research, Muellerstrasse 178, 13353 Berlin (Germany)], E-mail: hanns-joachim.weinmann@bayerhealthcare.com; Gerlach, Martin [Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin (Germany)], E-mail: martin.gerlach@ptb.de; Cibik, Levent [Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin (Germany)], E-mail: levent.cibik@ptb.de; Krumrey, Michael [Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin (Germany)], E-mail: michael.krumrey@ptb.de; Fratzscher, Daniel [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Fratzscher@ifg-adlershof.de; Rabe, Johannis [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Rabe@ifg-adlershof.de; Arkadiev, Vladimir [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Arkadiev@ifg-adlershof.de; Haschke, Michael [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Haschke@ifg-adlershof.de; Langhoff, Norbert [Institute for Scientific Instruments GmbH, Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: Langhoff@ifg-adlershof.de; Wedell, Reiner [Institut fuer angewandte Photonik e.V., Rudower Chaussee 29/31, 12489 Berlin (Germany)], E-mail: wedell-iap@ifg-adlershof.de (and others)

    2008-12-15

    Introduction: Computed tomography (CT) is a widespread and highly precise technique working in the energy range around 50-100 keV. For radiotherapy, however, the MeV energy range enables a better dose distribution. This gap between diagnosis and therapy can be overcome by the use of a modified CT machine in combination with heavy elements targeted to the tumour and used as photoelectric radiation enhancer. Materials and methods: The experimental setup consists of an X-ray optical module mounted at the exit of the X-ray tube of a clinical CT. The module converts the standard fan-shaped beam into a high intensity, monochromatized and focused beam. The radiation was characterized using an energy-dispersive detection system calibrated by synchrotron radiation and gel dosimetry. The photoelectric radiation enhancement for different elements was calculated and experimentally verified. Results: The X-ray optical module filters selectively the energy of the tungsten K{alpha}-emission line (59.3 keV) with a full width at half maximum (FWHM) of 5 keV and focused the radiation onto a focal spot which coincides with the isocentre of the gantry. This results in a steep dose gradient at the centre of rotation qualified for locoregional radiation therapy. The photon energy of the quasi-monochromatic radiation agrees with the energy range of maximal photoelectric dose enhancement for gadolinium and iodine. Conclusion: An additional X-ray optical module optimized for targeted therapy and photoelectric dose enhancement allows the combination of diagnosis and radiotherapy on a clinical CT.

  17. Three-dimensional monochromatic x-ray computed tomography using synchrotron radiation

    Science.gov (United States)

    Saito, Tsuneo; Kudo, Hiroyuki; Takeda, Tohoru; Itai, Yuji; Tokumori, Kenji; Toyofuku, Fukai; Hyodo, Kazuyuki; Ando, Masami; Nishimura, Katsuyuki; Uyama, Chikao

    1998-08-01

    We describe a technique of 3D computed tomography (3D CT) using monochromatic x rays generated by synchrotron radiation, which performs a direct reconstruction of a 3D volume image of an object from its cone-beam projections. For the development, we propose a practical scanning orbit of the x-ray source to obtain complete 3D information on an object, and its corresponding 3D image reconstruction algorithm. The validity and usefulness of the proposed scanning orbit and reconstruction algorithm were confirmed by computer simulation studies. Based on these investigations, we have developed a prototype 3D monochromatic x-ray CT using synchrotron radiation, which provides exact 3D reconstruction and material-selective imaging by using the K-edge energy subtraction technique.

  18. Summary of the dose assessment tool for diagnostic X-ray CT, WAZA-ARIv2

    International Nuclear Information System (INIS)

    Koba, Yosuke

    2016-01-01

    WAZA-ARIv2 is the web-based open system for the dose assessment for diagnostic X-ray CT. Amid growing interest about the dose assessment for medical exposure, WAZA-ARIv2 is opened to the public from January 2015 and attracts rising attention. Using WAZA-ARIv2 system, users can calculate exposure dose to consider patient's age and body shape, and can register the calculation results on WAZA-ARIv2 server for checking as histogram statistics. This paper reviews the background of development of WAZA-ARI system, the comparison of characteristics between WAZA-ARIv2 and other tools, and the calculation method of organ dose in WAZA-ARIv2. (author)

  19. X-ray CT-Derived Soil Characteristics Explain Varying Air, Water, and Solute Transport Properties across a Loamy Field

    DEFF Research Database (Denmark)

    Paradelo Pérez, Marcos; Katuwal, Sheela; Møldrup, Per

    2016-01-01

    -derived parameters by using a best subsets regression analysis. The regression coefficients improved using CTmatrix, limiting macroporosity, and genus density, while the best model for t0.05 used CTmatrix only. The scanning resolution and the time for soil structure development after mechanical activities could......The characterization of soil pore space geometry is important for explaining fluxes of air, water, and solutes through soil and understanding soil hydrogeochemical functions. X-ray computed tomography (CT) can be applied for this characterization, and in this study CT-derived parameters were used...... to explain water, air, and solute transport through soil. Forty-five soil columns (20 by 20 cm) were collected from an agricultural field in Estrup, Denmark, and subsequently scanned using a medical CT scanner. Nonreactive tracer leaching experiments were performed in the laboratory along with measurements...

  20. UGCT: New X-ray radiography and tomography facility

    International Nuclear Information System (INIS)

    Masschaele, B.C.; Cnudde, V.; Dierick, M.; Jacobs, P.; Hoorebeke, L. van; Vlassenbroeck, J.

    2007-01-01

    The UGCT (University Gent Computer Tomography) facility, a cooperation between the Radiation Physics research group and the Sedimentary Geology and Engineering Geology research group is a new CT facility providing a large range of scanning possibilities. Formerly a Skyscan 1072 was used to perform X-ray micro-CT scans at the UGCT facility and although this is a very powerful instrument, there were needs for a higher resolution and more flexibility. Therefore, the UCGT facility started the construction of a multidisciplinary micro-CT scanner inside a shielded room with a maximum flexibility of the set-up. The X-ray tube of this high-resolution CT scanner is a state-of-the-art open-type device with dual head: one head for high power micro-CT and one for sub-micro- or also called nano-CT. An important advantage of this scanner is that different detectors can be used to optimize the scanning conditions of the objects under investigation. The entire set-up is built on a large optical table to obtain the highest possible stability. Due to the flexible set-up and the powerful CT reconstruction software 'Octopus', it is possible to obtain the highest quality and the best signal-to-noise of the reconstructed images for each type of sample

  1. UGCT: New X-ray radiography and tomography facility

    Energy Technology Data Exchange (ETDEWEB)

    Masschaele, B.C. [Department of Subatomic and Radiation Physics, Ghent University, Proeftuinstraat 86, B-9000 Gent (Belgium)], E-mail: bert.masschaele@ugent.be; Cnudde, V. [Department of Geology and Soil Science, Ghent University, Krijgslaan 281, B-9000 Gent (Belgium); Dierick, M. [Department of Subatomic and Radiation Physics, Ghent University, Proeftuinstraat 86, B-9000 Gent (Belgium); Jacobs, P. [Department of Geology and Soil Science, Ghent University, Krijgslaan 281, B-9000 Gent (Belgium); Hoorebeke, L. van; Vlassenbroeck, J. [Department of Subatomic and Radiation Physics, Ghent University, Proeftuinstraat 86, B-9000 Gent (Belgium)

    2007-09-21

    The UGCT (University Gent Computer Tomography) facility, a cooperation between the Radiation Physics research group and the Sedimentary Geology and Engineering Geology research group is a new CT facility providing a large range of scanning possibilities. Formerly a Skyscan 1072 was used to perform X-ray micro-CT scans at the UGCT facility and although this is a very powerful instrument, there were needs for a higher resolution and more flexibility. Therefore, the UCGT facility started the construction of a multidisciplinary micro-CT scanner inside a shielded room with a maximum flexibility of the set-up. The X-ray tube of this high-resolution CT scanner is a state-of-the-art open-type device with dual head: one head for high power micro-CT and one for sub-micro- or also called nano-CT. An important advantage of this scanner is that different detectors can be used to optimize the scanning conditions of the objects under investigation. The entire set-up is built on a large optical table to obtain the highest possible stability. Due to the flexible set-up and the powerful CT reconstruction software 'Octopus', it is possible to obtain the highest quality and the best signal-to-noise of the reconstructed images for each type of sample.

  2. Timing of radiotherapy (RT) and chemotherapy (CT) following breast conserving surgery (BCS) for patients with node-positive breast cancer

    Energy Technology Data Exchange (ETDEWEB)

    Wallgren, A; Bernier, J; Gelber, RD; Roncadin, M; Joseph, D; Castiglione, M

    1995-07-01

    Purpose: To evaluate the effects of delaying RT until after different durations of initial courses of CT for pts with node-positive breast cancer. Methods and Materials: From 7/86-4/93, the IBCSG enrolled 1475 eligible node-positive pre- and perimenopausal pts in Trial 6 and 1212 eligible node-positive postmenopausal pts in Trial 7. Pts in Trial 6 were randomized to receive one of four treatments: (1) CMF x 6 cycles (1 cycle=28 days; cyclophosphamide 100 mg/m{sup 2} orally days 1-14; methotrexate 40 mg/m{sup 2} iv. days 1 and 8; 5-fluorouracil 600 mg/m{sup 2} iv days 1 and 8); (2) CMF x 6 cycles followed by single cycles at months 9, 12 and 15; (3) CMF x 3 cycles; (4) CMF x 3 followed by single cycles at months 6, 9 and 12. Pts in Trial 7 were randomized to receive one of four treatments: (5) tamoxifen (TAM: 20 mg daily) plus CMF x 3 cycles, months 1,2,3; (6) TAM plus CMF x 3 plus single cycles of CMF at months 9, 12 and 15; (7) TAM alone for five yrs; (8) TAM plus single cycles of CMF at months 9, 12 and 15. The randomizations were stratified according to type of surgery; 434 pts in Trial 6 (29%) and 285 pts in Trial 7 (24%) had breast conserving surgery and are the subjects of this report. Pts were randomized to receive RT to the conserved breast AFTER completing the initial block of CT if assigned; i.e. month 7 for arms 1 and 2; month 4 for arms 3 and 4; month 4 for arms 5 and 6; month 1 for arms 7 and 8. 24% of pts in Trial 6 and 27% of pts in Trial 7 had four or more axillary lymph nodes involved. Possible prognostic factors including age, estrogen receptor status, nodal status and tumor size were balanced across randomized groups. Four-year actuarial total failure rates (failure at any site), risks of developing distant metastases (DM) (at any time during observation), and overall survival (OS) were estimated using the Kaplan-Meier method. To avoid potential bias due to competing causes of failure, only patients who could be followed for at least 4 years

  3. Elemental distribution in ascending aortic after radiotherapy and chemotherapy by Low Energy X-ray Fluorescence spectroscopy

    Science.gov (United States)

    Mantuano, A.; Mota, C. L.; Pickler, A.; Sena, G.; Braz, D.; Salata, C.; de Almeida, C. E.; Costa, F. N.; Barroso, R. C.

    2018-05-01

    Breast cancer (BC) is the most frequent cancer and the leading cause of cancer-related mortality in women. The treatment techniques for the BC include chemotherapy (CT) and/or radiotherapy (RT) and can modify elementary the cell matrix by calcificating tissues due to biological and morphological changes. Also, treatments for BC induce cardiotoxicity and it is important to understand the mechanisms involved in order to prevent this late effect in treated breast cancer patients. The high incidence of cardiovascular mortality in breast cancer patients is partially credited to increased intimal and medial calcifications of the aorta. The aim of this work is to investigate the distibution of low atomic number elements such as Magnesium (Mg), due to its importance for the cardiac metabolism; iron (Fe), since BC treatment may be associated with oxidative stress; and Sodium (Na), that is extremely related to the damage of endothelial cells. An optimal technique to observe these changes in aorta tissue is soft X-ray FLuorescence that can provide elemental maps of these important elements. The results performed by Low Energy X-ray Fluorescence LEXRF analyses showed that when the tissue is submitted to treatments with CT and/or RT, some normal structures become disorganized, and consequently the intensity of elemental compounds can be changed. All the experiments were carried out at the TwinMic beamline at Elettra Synchrotron facility using as animal model Wistar rats in order to evaluate the distribution of Na, Mg and Fe in aorta walls of Wistar rats, after BC treatment. Simultaneous acquisition of LEXRF and attenuation coefficient maps suggest that the combined chemotherapy and radiotherapy caused more damage to the aortic tissue as compared to radiation therapy alone. These findings add an in-depth understanding of elemental lack or excess in the tissue and contribute to locate these changes.

  4. Results of a second questionnaire on pediatric X-ray examinations

    International Nuclear Information System (INIS)

    Ishikawa, Mitsuo; Kubota, Katsumi

    1999-01-01

    Radiographic conditions (x-ray tube voltage, irradiation time, etc.) differ among medical facilities. A second questionnaires was sent out to determine the current state of pediatric radiography and eventually provide reference materials for the standardization of radiographic parameters used in pediatric x-ray examinations. The questionnaire, which was sent in 1996, targeted 161 facilities that belong to the Society for Pediatric Radiological Technology. The objects of examination were the chest, upper airway, abdomen, hip joint, skull, and knee joint. The questionnaire investigated age (children of six months, three years, and seven years), type of x-ray generator, radiographic conditions, etc. Moreover, this time the items head x-ray computed tomography (X-CT) and abdominal X-CT were added. Completed questionnaires were received from 79 facilities, for a recovery of 49.1%. Compared with the previous investigation, inverter-type high-voltage assemblies increased, whereas twelve-peak high-voltage assemblies decreased. The focal spot of the x-ray tube assembly was smaller, and maximum anode heat content had increased. CR systems and green luminescence-orthochromatic systems were increasingly being used as image receptors. Advances in x-ray generators and the image receptor systems created reductions in the time required for radiography, which was possible at as short a time as 1 msec. Moreover, image quality was also improved by these advances. However, no major changes were observed in the conditions of radiography. (author)

  5. Multiple energy computed tomography with monochromatic x rays from the NSLS

    International Nuclear Information System (INIS)

    Dilmanian, F.A.; Nachaliel, E.; Garrett, R.F.; Thomlinson, W.C.; Chapman, L.D.; Moulin, H.R.; Oversluizen, T.; Rarback, H.M.; Rivers, M.; Spanne, P.; Thompson, A.C.; Zeman, H.D.

    1991-01-01

    We used monochromatic x rays from the X17 superconducting wiggler beamline at the National Synchrotron Light Source (NSLS), Brookhaven National Laboratory, for dual-energy quantitative computed tomography (CT) of a 27 mm-diameter phantom containing solutions of different KOH concentrations in cylindrical holes of 5-mm diameter. The CT configuration was a fixed horizontal fan-shaped beam of 1.5 mm height and 30 mm width, and a subject rotating around a vertical axis. The transmitted x rays were detected by a linear-array Si(Li) detector with 120 elements of 0.25 mm width each. We used a two-crystal Bragg-Bragg fixed-exit monochromator with Si crystals. Dual photon absorptiometry (DPA) CT data were taken at 20 and 38 keV. The reconstructed phantom images show the potential of the system for quantitative CT

  6. Utility of three-dimensional helical CT in the diagnosis of breast cancer

    International Nuclear Information System (INIS)

    Maeda, Yoshiaki; Hata, Yoshinobu; Matsuoka, Shinnichi; Nakajima, Nobuhisa; Ito, Toichi; Osada, Tadahiro; Sano, Fumio

    2004-01-01

    Although utility of three-dimensional (3D) helical CT for preoperative examination of breast cancer has been discussed, the accuracy of the helical CT in diagnosing breast cancer has not been fully evaluated. In this study 56 malignant and 28 benign breast tumors were evaluated preoperatively with 3D-helical CT, and their imaging results were compared with pathological findings of surgical specimens. Helical CT identified the presence of malignancy in 54 out of the 56 cancer cases tested and the sensitivity and specificity in distinguishing between malignant and benign tumors were 82% and 57%, respectively. The sensitivity and specificity in diagnosing the presence of metastatic axillary lymph nodes using helical CT were 70% and 80%, respectively. The sensitivity and specificity in diagnosing the presence of extensive intraductal component (EIC) using helical CT were 71% and 86%, respectively. Helical CT visualized all of the tumors in multifocal breast cancer cases. In conclusion, 3D-helical CT is a useful modality for preoperative examination of breast cancer, especially for assessing axillary lymph node status, and EIC, and will be helpful for conducting sentinel lymph node biopsy (SNLB) and breast-conserving surgery. (author)

  7. SU-D-206-06: Task-Specific Optimization of Scintillator Thickness for CMOS-Detector Based Cone-Beam Breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Vedantham, S; Shrestha, S; Shi, L; Vijayaraghavan, G; Karellas, A [University of Massachusetts Medical School, Worcester, MA (United States)

    2016-06-15

    Purpose: To optimize the cesium iodide (CsI:Tl) scintillator thickness in a complimentary metal-oxide semiconductor (CMOS)-based detector for use in dedicated cone-beam breast CT. Methods: The imaging task considered was the detection of a microcalcification cluster comprising six 220µm diameter calcium carbonate spheres, arranged in the form of a regular pentagon with 2 mm spacing on its sides and a central calcification, similar to that in ACR-recommended mammography accreditation phantom, at a mean glandular dose of 4.5 mGy. Generalized parallel-cascades based linear systems analysis was used to determine Fourier-domain image quality metrics in reconstructed object space, from which the detectability index inclusive of anatomical noise was determined for a non-prewhitening numerical observer. For 300 projections over 2π, magnification-associated focal-spot blur, Monte Carlo derived x-ray scatter, K-fluorescent emission and reabsorption within CsI:Tl, CsI:Tl quantum efficiency and optical blur, fiberoptic plate transmission efficiency and blur, CMOS quantum efficiency, pixel aperture function and additive noise, and filtered back-projection to isotropic 105µm voxel pitch with bilinear interpolation were modeled. Imaging geometry of a clinical prototype breast CT system, a 60 kV Cu/Al filtered x-ray spectrum from 0.3 mm focal spot incident on a 14 cm diameter semi-ellipsoidal breast were used to determine the detectability index for 300–600 µm thick (75µm increments) CsI:Tl. The CsI:Tl thickness that maximized the detectability index was considered optimal. Results: The limiting resolution (10% modulation transfer function, MTF) progressively decreased with increasing CsI:Tl thickness. The zero-frequency detective quantum efficiency, DQE(0), in projection space increased with increasing CsI:Tl thickness. The maximum detectability index was achieved with 525µm thick CsI:Tl scintillator. Reduced MTF at mid-to-high frequencies for 600µm thick CsI:Tl lowered

  8. SU-F-I-14: 3D Breast Digital Phantom for XACT Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Tang, S; Laaroussi, R; Chen, J; Samant, P; Xiang, L [University of Oklahoma, Norman, OK (United States); Chen, Y; Ahmad, S [University of Oklahoma Health Sciences Center, Oklahoma City, OK (United States); Yang, K [Massachusetts General Hospital, Boston, MA (United States)

    2016-06-15

    Purpose: The X-ray induced acoustic computed tomography (XACT) is a new imaging modality which combines X-ray contrast and high ultrasonic resolution in a single modality. Using XACT in breast imaging, a 3D breast volume can be imaged by only one pulsed X-ray radiation, which could dramatically reduce the imaging dose for patients undergoing breast cancer screening and diagnosis. A 3D digital phantom that contains both X-ray properties and acoustic properties of different tissue types is indeed needed for developing and optimizing the XACT system. The purpose of this study is to offer a realistic breast digital phantom as a valuable tool for improving breast XACT imaging techniques and potentially leading to better diagnostic outcomes. Methods: A series of breast CT images along the coronal plane from a patient who has breast calcifications are used as the source images. A HU value based segmentation algorithm is employed to identify breast tissues in five categories, namely the skin tissue, fat tissue, glandular tissue, chest bone and calcifications. For each pixel, the dose related parameters, such as material components and density, and acoustic related parameters, such as frequency-dependent acoustic attenuation coefficient and bandwidth, are assigned based on tissue types. Meanwhile, other parameters which are used in sound propagation, including the sound speed, thermal expansion coefficient, and heat capacity are also assigned to each tissue. Results: A series of 2D tissue type image is acquired first and the 3D digital breast phantom is obtained by using commercial 3D reconstruction software. When giving specific settings including dose depositions and ultrasound center frequency, the X-ray induced initial pressure rise can be calculated accordingly. Conclusion: The proposed 3D breast digital phantom represents a realistic breast anatomic structure and provides a valuable tool for developing and evaluating the system performance for XACT.

  9. Characterization of test specimens produced in reduced size for X-ray microtomography (µ-CT tests

    Directory of Open Access Journals (Sweden)

    E. E. BERNARDES

    Full Text Available Abstract The need to use reduced sample sizes, in order to attain improved spatial resolution in (µ-CT tests applied in Portland cement composites, makes researchers perform the fractionation of materials to obtain samples with dimensions compatible with the capacity of the scanning equipment, which might cause alterations in the microstructure under analysis. Therefore, a test specimen (TS with dimensions compatible with the scanning capacity of a microtomography system that operates with an X-ray tube and voltage ranging from 20 to 100 kV was proposed. Axial compression strength tests were made and their total porosity was assessed by an apparent density and solid fraction density ratio, which were obtained by means of mercury and helium pycnometry and µ-CT technique, respectively. The adoption of that TS has shown to be viable for providing a sample with a higher level of representation.

  10. Development of an MRI/x-ray/ultrasound compatible marker for pre-operative breast tumour localization

    International Nuclear Information System (INIS)

    Li Yangmei; Wang, Jianxiong; Holloway, Claire; Plewes, Donald B

    2005-01-01

    This paper describes an in vitro investigation into the composition, structure and development of an magnetic resonance imaging (MRI), ultrasound (US) and x-ray imaging compatible marker for breast tumour localization. The marker is composed of 0.4-0.6 mm glass and iron-containing aluminium microspheres suspended in a gelatin matrix. The final form of the marker is a cylindrical shape 7 mm long with 2.05 mm diameter to facilitate delivery through a 12 gauge biopsy needle. To get optimal reflectivity for the US contrast, the glass microsphere concentration was found to be 40% by weight. US contrast is independent of marker orientation and the cylindrical shape made its US signal appearance distinctive thus ensuring confident identification. To control the MRI contrast, iron content was varied to generate a clear and local susceptibility signal void to reflect the marker position. Optimal iron content was found to be 52 μg iron which produced a clear signal void in spoiled gradient recalled MR images. The appearance of the susceptibility artefact is determined by the marker's shape, orientation and echo time. The final marker produces a dark artefact in MRI while appears as a clear hyperintense structure with acoustic shadowing in US images. The x-ray image showed the marker as a radio-opaque structure. This in vitro study demonstrates that the marker forms an alternative to traditional wire localization currently used for breast surgical procedures and creates new opportunities for US guided surgical procedures

  11. Refinement of clinical X-ray computed tomography (CT) scans containing metal implants.

    Science.gov (United States)

    Treece, Graham

    2017-03-01

    X-ray computed tomography (CT) data contains artefacts from many sources, with sufficient prominence to affect diagnostic utility when metal is present in the scans. These artefacts can be reduced, usually by the removal and in-filling of any sinogram data which has been affected by metal, and several such techniques have been proposed. Most of them are prone to introducing new artefacts into the CT data or may take a long time to correct the data. It is the purpose of this paper to introduce a new technique which is fast, yet can effectively remove most artefacts without introducing significant new ones. The new metal artefact reduction technique (RMAR) consists of an iterative refinement of the CT data by alternately forward- and back-projecting the part of the reconstruction near to metal. The forward-projection is corrected by making use of a prior derived from the reconstructed data which is independently estimated for each projection angle, and smoothed using a newly developed Bitonic filter. The new technique is compared with previously published (LI, NMAR, MDT) and commercial (O-MAR, IMAR) alternatives, quantitatively on phantom data, and qualitatively on a selection of clinical scans, mostly of the hip. The phantom data is from two recently published studies, enabling direct comparison with previous results. The results show an increased reduction of artefacts on the four phantom data sets tested. On two of the phantom data sets, RMAR is significantly better (pCT data, RMAR can correct each image in 3-8s, which is more than one hundred times faster than MDT. The new technique is demonstrated to have performance at least as good as MDT, with both out-performing other approaches. However, it is much faster then the latter technique, and shows better preservation of data very close to metal. Copyright © 2017 Elsevier Ltd. All rights reserved.

  12. X-ray imaging with sub-micron resolution using large-area photon counting detectors Timepix

    Science.gov (United States)

    Dudak, J.; Karch, J.; Holcova, K.; Zemlicka, J.

    2017-12-01

    As X-ray micro-CT became a popular tool for scientific purposes a number of commercially available CT systems have emerged on the market. Micro-CT systems have, therefore, become widely accessible and the number of research laboratories using them constantly increases. However, even when CT scans with spatial resolution of several micrometers can be performed routinely, data acquisition with sub-micron precision remains a complicated task. Issues come mostly from prolongation of the scan time inevitably connected with the use of nano-focus X-ray sources. Long exposure time increases the noise level in the CT projections. Furthermore, considering the sub-micron resolution even effects like source-spot drift, rotation stage wobble or thermal expansion become significant and can negatively affect the data. The use of dark-current free photon counting detectors as X-ray cameras for such applications can limit the issue of increased image noise in the data, however the mechanical stability of the whole system still remains a problem and has to be considered. In this work we evaluate the performance of a micro-CT system equipped with nano-focus X-ray tube and a large area photon counting detector Timepix for scans with effective pixel size bellow one micrometer.

  13. X-ray computed tomography

    International Nuclear Information System (INIS)

    Kalender, Willi A

    2006-01-01

    X-ray computed tomography (CT), introduced into clinical practice in 1972, was the first of the modern slice-imaging modalities. To reconstruct images mathematically from measured data and to display and to archive them in digital form was a novelty then and is commonplace today. CT has shown a steady upward trend with respect to technology, performance and clinical use independent of predictions and expert assessments which forecast in the 1980s that it would be completely replaced by magnetic resonance imaging. CT not only survived but exhibited a true renaissance due to the introduction of spiral scanning which meant the transition from slice-by-slice imaging to true volume imaging. Complemented by the introduction of array detector technology in the 1990s, CT today allows imaging of whole organs or the whole body in 5 to 20 s with sub-millimetre isotropic resolution. This review of CT will proceed in chronological order focussing on technology, image quality and clinical applications. In its final part it will also briefly allude to novel uses of CT such as dual-source CT, C-arm flat-panel-detector CT and micro-CT. At present CT possibly exhibits a higher innovation rate than ever before. In consequence the topical and most recent developments will receive the greatest attention. (review)

  14. Investigation of optimal scanning protocol for X-ray computed tomography polymer gel dosimetry

    Energy Technology Data Exchange (ETDEWEB)

    Sellakumar, P. [Bangalore Institute of Oncology, 44-45/2, II Cross, RRMR Extension, Bangalore 560 027 (India)], E-mail: psellakumar@rediffmail.com; James Jebaseelan Samuel, E. [School of Science and Humanities, VIT University, Vellore 632 014 (India); Supe, Sanjay S. [Department of Radiation Physics, Kidwai Memorial Institute of Oncology, Hosur Road, Bangalore 560 027 (India)

    2007-11-15

    X-ray computed tomography is one of the potential tool used to evaluate the polymer gel dosimeters in three dimensions. The purpose of this study is to investigate the factors which affect the image noise for X-ray CT polymer gel dosimetry. A cylindrical water filled phantom was imaged with single slice Siemens Somatom Emotion CT scanner. The imaging parameters like tube voltage, tube current, slice scan time, slice thickness and reconstruction algorithm were varied independently to study the dependence of noise on each other. Reductions of noise with number of images to be averaged and spatial uniformity of the image were also investigated. Normoxic polymer gel PAGAT was manufactured and irradiated using Siemens Primus linear accelerator. The radiation induced change in CT number was evaluated using X-ray CT scanner. From this study it is clear that image noise is reduced with increase in tube voltage, tube current, slice scan time, slice thickness and also reduced with increasing the number of images averaged. However to reduce the tube load and total scan time, it was concluded that tube voltage of 130 kV, tube current of 200 mA, scan time of 1.5 s, slice thickness of 3 mm for high dose gradient and 5 mm for low dose gradient were optimal scanning protocols for this scanner. Optimum number of images to be averaged was concluded to be 25 for X-ray CT polymer gel dosimetry. Choice of reconstruction algorithm was also critical. From the study it is also clear that CT number increase with imaging tube voltage and shows the energy dependency of polymer gel dosimeter. Hence for evaluation of polymer gel dosimeters with X-ray CT scanner needs the optimization of scanning protocols to reduce the image noise.

  15. TU-F-CAMPUS-I-02: Validation of a CT X-Ray Source Characterization Technique for Dose Computation Using An Anthropomorphic Thorax Phantom

    Energy Technology Data Exchange (ETDEWEB)

    Sommerville, M; Tambasco, M [San Diego State University, San Diego, CA (United States); Poirier, Y [CancerCare Manitoba, Winnipeg, MB (Canada)

    2015-06-15

    Purpose: To experimentally validate a rotational kV x-ray source characterization technique by computing CT dose in an anthropomorphic thorax phantom using an in-house dose computation algorithm (kVDoseCalc). Methods: The lateral variation in incident energy spectra of a GE Optima big bore CT scanner was found by measuring the HVL along the internal, full bow-tie filter axis. The HVL and kVp were used to generate the x-ray spectra using Spektr software, while beam fluence was derived by dividing the integral product of the spectra and in-air mass-energy absorption coefficients by in-air dose measurements along the bow-tie filter axis. Beams produced by the GE Optima scanner were modeled at 80 and 140 kVp tube settings. kVDoseCalc calculates dose by solving the linear Boltzmann transport equation using a combination of deterministic and stochastic methods. Relative doses in an anthropomorphic thorax phantom (E2E SBRT Phantom) irradiated by the GE Optima scanner were measured using a (0.015 cc) PTW Freiburg ionization chamber, and compared to computations from kVDoseCalc. Results: The agreement in relative dose between dose computation and measurement for points of interest (POIs) within the primary path of the beam was within experimental uncertainty for both energies, however points outside the primary beam were not. The average absolute percent difference for POIs within the primary path of the beam was 1.37% and 5.16% for 80 and 140 kVp, respectively. The minimum and maximum absolute percent difference for both energies and all POIs within the primary path of the beam was 0.151% and 6.41%, respectively. Conclusion: The CT x-ray source characterization technique based on HVL measurements and kVp can be used to accurately compute CT dose in an anthropomorphic thorax phantom.

  16. Dual-time FDG-PET/CT in patients with potential breast cancer recurrence

    DEFF Research Database (Denmark)

    Baun, Christina; Falch Braas, Kirsten; Gerke, Oke

    Dual-time FDG-PET/CT in patients with potential breast cancer recurrence: head-to-head comparison with CT and bonescintigraphy......Dual-time FDG-PET/CT in patients with potential breast cancer recurrence: head-to-head comparison with CT and bonescintigraphy...

  17. X-ray stereotactic lesion localization in conjunction with dedicated scintimammography

    Energy Technology Data Exchange (ETDEWEB)

    Mitali J. More; Deepa Narayanan; Patricia J. Goodale; Stanislaw Majewski; Benjamin Welch; Randolph Wojcik; Mark Williams

    2003-10-01

    We are developing a dual modality system that combines digital X-ray mammography with gamma emission scintigraphy on an upright mammography gantry. The breast is held under mild compression by a support structure that is independent of the detectors. The X-ray source and detectors can be rotated around a fixed rotation axis permitting multiple views of the breast with fixed compression. Two such views can be combined as a stereotactic pair to obtain the three-dimensional location of breast lesions. Information about the location of the lesion within the breast permits corrections for attenuation and detector spatial resolution, resulting in more accurate estimation of the true lesion-to-background concentration ratio, based on the image lesion-to-background counts ratio. In this paper, we describe the model used to make these corrections, and present the results of the phantom experiments designed to test the accuracy of our calculations.

  18. Ultra-fast x-ray tomography for multi-phase flow interface dynamic studies

    International Nuclear Information System (INIS)

    Misawa, M.; Ichikawa, N.; Akai, M.; Tiseanu, I.; Prasser, H.-M.

    2003-01-01

    The present paper describes the concept of a fast scanning X-ray tomograph, the hardware development, and measurement results of gas-liquid two-phase flow in a vertical pipe. The device uses 18 pulsed X-ray sources activated in a successive order. In this way, a complete set of 18 independent projections of the object is obtained within 38 ms, i.e. the measuring rate is about 250 frames per second. Finally, to evaluate the measurement capability of the fast X-ray CT, a wire-mesh sensor was installed in the flow loop and both systems were operated for the same two-phase flow simultaneously. Comparison of the time series of the cross section averaged void fraction from both systems showed sufficient agreement for slug flow at large void fractions, while the fast CT underestimated the void fraction of bubbly flow especially in low void fraction range. For the wire-mesh sensor, coerced deformation of slug bubble interface was found. Further hardware improvement is in progress to achieve better resolution with the fast X-ray CT scanner. (orig.)

  19. TU-EF-207-04: Advances in Detector Technology for Breast Tomosynthesis

    Energy Technology Data Exchange (ETDEWEB)

    Zhao, W. [SUNY Stony Brook (United States)

    2015-06-15

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation.

  20. TU-EF-207-04: Advances in Detector Technology for Breast Tomosynthesis

    International Nuclear Information System (INIS)

    Zhao, W.

    2015-01-01

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation

  1. Analysis of iterative region-of-interest image reconstruction for x-ray computed tomography

    Science.gov (United States)

    Sidky, Emil Y.; Kraemer, David N.; Roth, Erin G.; Ullberg, Christer; Reiser, Ingrid S.; Pan, Xiaochuan

    2014-01-01

    Abstract. One of the challenges for iterative image reconstruction (IIR) is that such algorithms solve an imaging model implicitly, requiring a complete representation of the scanned subject within the viewing domain of the scanner. This requirement can place a prohibitively high computational burden for IIR applied to x-ray computed tomography (CT), especially when high-resolution tomographic volumes are required. In this work, we aim to develop an IIR algorithm for direct region-of-interest (ROI) image reconstruction. The proposed class of IIR algorithms is based on an optimization problem that incorporates a data fidelity term, which compares a derivative of the estimated data with the available projection data. In order to characterize this optimization problem, we apply it to computer-simulated two-dimensional fan-beam CT data, using both ideal noiseless data and realistic data containing a level of noise comparable to that of the breast CT application. The proposed method is demonstrated for both complete field-of-view and ROI imaging. To demonstrate the potential utility of the proposed ROI imaging method, it is applied to actual CT scanner data. PMID:25685824

  2. X-ray phase-contrast tomosynthesis for improved breast tissue discrimination

    International Nuclear Information System (INIS)

    Schleede, Simone; Bech, Martin; Grandl, Susanne; Sztrókay, Aniko; Herzen, Julia; Mayr, Doris; Stockmar, Marco; Potdevin, Guillaume

    2014-01-01

    Purpose: Attenuation-based tomosynthesis has proven to successfully resolve the glandular tissue overlap present in mammography. However, the ability of tomosynthesis to differentiate tumorous and glandular tissue remains limited, due to the small differences in X-ray attenuation in breast tissue. One possibility to overcome this limitation and to further increase the diagnostic value of tomosynthesis exams, is the application of recently developed grating-based phase-contrast methods, which provide complementary information on the phase shift and the local scattering power of the sample. In this study, we report on first phase-contrast breast tomosynthesis results of a mastectomy sample slice with an invasive ductal carcinoma. Material and methods: A slice of a mastectomy sample with histologically proven invasive ductal cancer was imaged at the synchrotron radiation source ESRF (Grenoble, France). We used a two-grating interferometer setup at the ninth fractional Talbot distance and with an X-ray energy of 23 keV. In grating interferometry absorption, differential phase, and scattering images are recorded simultaneously. The tomosynthesis scan comprises 61 projections. Multimodal tomosynthesis results were reconstructed using a standard filtered back-projection approach. Our findings are supported by a comparison of tomographic views to histopathology. Results: Phase-contrast tomosynthesis combines the advantage of improved soft-tissue discrimination in phase-contrast imaging with the ability of tomosynthesis to provide a third dimension so that improved feature visibility is not hampered by superposition artifacts. Our results indicate superior diagnostic value due to the depth resolution supplied in tomosynthesis imaging; a region of necrotic tissue that is obscured in a projection image can clearly be depicted in one single tomosynthesis slice. Compared to absorption tomosynthesis alone, soft tissue contrast is significantly enhanced in phase

  3. X-ray phase-contrast tomosynthesis for improved breast tissue discrimination.

    Science.gov (United States)

    Schleede, Simone; Bech, Martin; Grandl, Susanne; Sztrókay, Aniko; Herzen, Julia; Mayr, Doris; Stockmar, Marco; Potdevin, Guillaume; Zanette, Irene; Rack, Alexander; Weitkamp, Timm; Pfeiffer, Franz

    2014-03-01

    Attenuation-based tomosynthesis has proven to successfully resolve the glandular tissue overlap present in mammography. However, the ability of tomosynthesis to differentiate tumorous and glandular tissue remains limited, due to the small differences in X-ray attenuation in breast tissue. One possibility to overcome this limitation and to further increase the diagnostic value of tomosynthesis exams, is the application of recently developed grating-based phase-contrast methods, which provide complementary information on the phase shift and the local scattering power of the sample. In this study, we report on first phase-contrast breast tomosynthesis results of a mastectomy sample slice with an invasive ductal carcinoma. A slice of a mastectomy sample with histologically proven invasive ductal cancer was imaged at the synchrotron radiation source ESRF (Grenoble, France). We used a two-grating interferometer setup at the ninth fractional Talbot distance and with an X-ray energy of 23 keV. In grating interferometry absorption, differential phase, and scattering images are recorded simultaneously. The tomosynthesis scan comprises 61 projections. Multimodal tomosynthesis results were reconstructed using a standard filtered back-projection approach. Our findings are supported by a comparison of tomographic views to histopathology. Phase-contrast tomosynthesis combines the advantage of improved soft-tissue discrimination in phase-contrast imaging with the ability of tomosynthesis to provide a third dimension so that improved feature visibility is not hampered by superposition artifacts. Our results indicate superior diagnostic value due to the depth resolution supplied in tomosynthesis imaging; a region of necrotic tissue that is obscured in a projection image can clearly be depicted in one single tomosynthesis slice. Compared to absorption tomosynthesis alone, soft tissue contrast is significantly enhanced in phase-contrast tomosynthesis views, where fibrous structures

  4. FDG-PET/CT in the diagnosis of recurrent breast cancer

    International Nuclear Information System (INIS)

    Murakami, Ryusuke; Kumita, Shin-ichiro; Yoshida, Tamiko; Ishihara, Keiichi; Kiriyama, Tomonari; Hakozaki, Kenta; Yanagihara, Keiko; Lida, Shinya; Tsuchiya, Shin-ichi

    2012-01-01

    Background. An advantage of PET/CT has been demonstrated for diagnosis of several tumor entities. In patients with breast cancer, early diagnosis and accurate restaging of recurrence after surgery is important for selection of the most appropriate therapeutic strategy. Purpose. To evaluate the accuracy of integrated positron emission tomography and computed tomography (PET/CT) using 18F-fluorodeoxyglucose (FDG), for follow-up of patients with suspected recurrent breast cancer. Material and Methods. Forty-seven patients with suspected recurrent breast cancer underwent PET/CT. The PET and PET/CT images were interpreted without knowledge of the results of other diagnostic modalities, and compared with each other with reference to the final diagnosis. Results. Twenty-five (53%) patients suffered tumor recurrence. The overall sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and accuracy of PET/CT were 96%, 91%, 92%, 95%, and 94%, respectively. In comparison with PET, PET/CT had a higher sensitivity and accuracy (96% vs. 80% and 94% vs. 81%, respectively). The difference in diagnostic accuracy between PET/CT and PET was significant (P < 0.05). Conclusion. The present findings indicate that PET/CT is an accurate, sensitive and reliable modality for screening and detection of breast cancer recurrence. PET/CT appears to be an effective surveillance tool, as it is able to cover the whole body in a single procedure and shows good performance

  5. Cancer among medical diagnostic x-ray workers in China

    International Nuclear Information System (INIS)

    Wang, J.X.; Boice, J.D. Jr.; Li, B.X.; Zhang, J.Y.; Fraumeni, J.F. Jr.

    1988-01-01

    Cancer incidence among 27,011 diagnostic x-ray workers was compared to that of 25,782 other medical specialists employed between 1950 and 1980 in China. X-ray workers had a 50% higher risk of developing cancer than the other specialists [relative risk (RR) = 1.5; 95% CI = 1.3-1.7]. Leukemia was strongly linked to radiation work (RR = 3.5, n = 30). Cancers of the breast (RR = 1.4, n = 11), thyroid (RR = 2.1, n = 7), and skin (RR = 1.5, n = 6) were increased among x-ray workers employed for 10 or more years. High risks of cancers of the esophagus (RR = 3.5, n = 15) and liver (RR = 2.4, n = 48) were not consistent with a radiation effect since risk did not vary by duration of employment. This finding suggested that some differences might exist between groups of hospital workers in social class, alcohol intake, dietary habits, and other risk factors. No excess lung cancer (RR = 0.9, n = 22) or multiple myeloma (n = 0) was observed. Significant excesses of leukemia and cancers of the breast and thyroid occurred among x-ray workers first employed prior to 1960 when radiation exposures in China were high. In fact, it was not uncommon for employees to be given time off from x-ray work because their wbc count was severely depressed. These data indicated that repeated exposure to x-rays over many years can increase the risk of leukemia and several other tumors but apparently not that of lung cancer

  6. Breast tumor segmentation in high resolution x-ray phase contrast analyzer based computed tomography

    Energy Technology Data Exchange (ETDEWEB)

    Brun, E., E-mail: emmanuel.brun@esrf.fr [European Synchrotron Radiation Facility (ESRF), Grenoble 380000, France and Department of Physics, Ludwig-Maximilians University, Garching 85748 (Germany); Grandl, S.; Sztrókay-Gaul, A.; Gasilov, S. [Institute for Clinical Radiology, Ludwig-Maximilians-University Hospital Munich, 81377 Munich (Germany); Barbone, G. [Department of Physics, Harvard University, Cambridge, Massachusetts 02138 (United States); Mittone, A.; Coan, P. [Department of Physics, Ludwig-Maximilians University, Garching 85748, Germany and Institute for Clinical Radiology, Ludwig-Maximilians-University Hospital Munich, 81377 Munich (Germany); Bravin, A. [European Synchrotron Radiation Facility (ESRF), Grenoble 380000 (France)

    2014-11-01

    Purpose: Phase contrast computed tomography has emerged as an imaging method, which is able to outperform present day clinical mammography in breast tumor visualization while maintaining an equivalent average dose. To this day, no segmentation technique takes into account the specificity of the phase contrast signal. In this study, the authors propose a new mathematical framework for human-guided breast tumor segmentation. This method has been applied to high-resolution images of excised human organs, each of several gigabytes. Methods: The authors present a segmentation procedure based on the viscous watershed transform and demonstrate the efficacy of this method on analyzer based phase contrast images. The segmentation of tumors inside two full human breasts is then shown as an example of this procedure’s possible applications. Results: A correct and precise identification of the tumor boundaries was obtained and confirmed by manual contouring performed independently by four experienced radiologists. Conclusions: The authors demonstrate that applying the watershed viscous transform allows them to perform the segmentation of tumors in high-resolution x-ray analyzer based phase contrast breast computed tomography images. Combining the additional information provided by the segmentation procedure with the already high definition of morphological details and tissue boundaries offered by phase contrast imaging techniques, will represent a valuable multistep procedure to be used in future medical diagnostic applications.

  7. Preoperative PET/CT in early-stage breast cancer

    DEFF Research Database (Denmark)

    Bernsdorf, M; Berthelsen, A K; Timmermans-Wielenga, Vera

    2012-01-01

    The aim of this study was to assess the diagnostic and therapeutic impact of preoperative positron emission tomography and computed tomography (PET/CT) in the initial staging of patients with early-stage breast cancer.......The aim of this study was to assess the diagnostic and therapeutic impact of preoperative positron emission tomography and computed tomography (PET/CT) in the initial staging of patients with early-stage breast cancer....

  8. Clinical Significance of Focal Breast Lesions Incidentally Identified by 18F-FDG PET/CT

    International Nuclear Information System (INIS)

    Cho, Young Seok; Choi, Joon Young; Lee, Su Jin; Hyun, Seung Hyup; Lee, Ji Young; Choi, Yong; Choe, Yearn Seong; Lee, Kyung Han; Kim, Byung Tae

    2008-01-01

    We evaluated the incidence and malignant risk of focal breast lesions incidentally detected by 18 F-FDG PET/CT. Various PET/CT findings of the breast lesions were also analyzed to improve the differentiation between benign from malignant focal breast lesions. The subjects were 3,768 consecutive 18 F-FDG PET/CT exams performed in adult females without a history of breast cancer. A focal breast lesion was defined as a focal 18 F-FDG uptake or a focal nodular lesion on CT image irrespective of 18 F-FDG uptake in the breasts. The maximum SUV and CT pattern of focal breast lesions were evaluated, and were compared with final diagnosis. The incidence of focal breast lesions on PET/CT in adult female subjects was 1.4% (58 lesions in 53 subjects). In finally confirmed 53 lesions of 48 subjects, 11 lesions of 8 subjects (20.8%) were proven to be malignant. When the PET/CT patterns suggesting benignancy (maximum attenuation value > 75 HU or 20) were added as diagnostic criteria of PET/CT to differentiate benign from malignant breast lesions along with maximum SUV, the area under ROC curve of PET/CT was significantly increased compared with maximum SUV alone (0.680±0.093 vs. 0.786±0.076, p 18 F-FDG PET/CT is not low, deserving further diagnostic confirmation. Image interpretation considering both 18 F-FDG uptake and PET/CT pattern may be helpful to improve the differentiation from malignant and benign focal breast lesion

  9. Evaluation of a breast software model for 2D and 3D X-ray imaging studies of the breast.

    Science.gov (United States)

    Baneva, Yanka; Bliznakova, Kristina; Cockmartin, Lesley; Marinov, Stoyko; Buliev, Ivan; Mettivier, Giovanni; Bosmans, Hilde; Russo, Paolo; Marshall, Nicholas; Bliznakov, Zhivko

    2017-09-01

    In X-ray imaging, test objects reproducing breast anatomy characteristics are realized to optimize issues such as image processing or reconstruction, lesion detection performance, image quality and radiation induced detriment. Recently, a physical phantom with a structured background has been introduced for both 2D mammography and breast tomosynthesis. A software version of this phantom and a few related versions are now available and a comparison between these 3D software phantoms and the physical phantom will be presented. The software breast phantom simulates a semi-cylindrical container filled with spherical beads of different diameters. Four computational breast phantoms were generated with a dedicated software application and for two of these, physical phantoms are also available and they are used for the side by side comparison. Planar projections in mammography and tomosynthesis were simulated under identical incident air kerma conditions. Tomosynthesis slices were reconstructed with an in-house developed reconstruction software. In addition to a visual comparison, parameters like fractal dimension, power law exponent β and second order statistics (skewness, kurtosis) of planar projections and tomosynthesis reconstructed images were compared. Visually, an excellent agreement between simulated and real planar and tomosynthesis images is observed. The comparison shows also an overall very good agreement between parameters evaluated from simulated and experimental images. The computational breast phantoms showed a close match with their physical versions. The detailed mathematical analysis of the images confirms the agreement between real and simulated 2D mammography and tomosynthesis images. The software phantom is ready for optimization purpose and extrapolation of the phantom to other breast imaging techniques. Copyright © 2017 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

  10. Analysis about X-ray and CT images of pleural mesothelioma case of death. Examination about 2003 mesothelioma case of death 878 examples

    International Nuclear Information System (INIS)

    Kato, Katsuya; Kanazawa, Susumu; Kishimoto, Takumi; Genba, Kenichi; Aoe, Keisuke; Takeshima, Yukio; Inai, Kouki

    2008-01-01

    We investigated a clinical record and images for 212 examples in 878 examples diagnosed as mesothelioma in the death votes of 2003. CT images demonstrated pleural plaque to be obtained 42.0%, but in chest X-ray films only 9.8% demonstrated pleural plaque. For examination of CT findings of 117 examples, cases to present extensive irregular findings of pleura were 81.2%. Cases we made the ''no irregularity'' and ''slightness irregularity'' were 18.8%. (author)

  11. Multi-scale characteristics of coal structure by x-ray computed tomography (x-ray CT), scanning electron microscope (SEM) and mercury intrusion porosimetry (MIP)

    Science.gov (United States)

    Cai, Ting-ting; Feng, Zeng-chao; Zhou, Dong

    2018-02-01

    It is of great benefit to study the material and structural heterogeneity of coal for better understanding the coalbed methane (CBM) storage and enrichment. In this paper, multi-scale X-ray computed tomography (CT), scanning electron microscope (SEM) and mercury intrusion porosimetry (MIP) at multi scales were conducted to thoroughly study the material distribution, heterogeneity, pore development, porosity and permeability of coal. It is suitable and reasonable to divide the testing samples into three structural categories by average density and heterogeneity degree, and the meso structure in the three categories accords with the morphology on SEM images. The pore size distribution and pore development of each subsample cannot be correspondingly related to their respective structure category or morphology due to different observation scales, while the macro pore size development, accumulated macro pore volume and macro pores porosity accord with the meso structure category and morphology information by CT and SEM at the same scale very well. Given the effect of macro pores on permeability and the contribution of micro pores to CBM storage capacity, reservoirs with developed micro pores and macro pores may be the most suitable coal reservoir for CBM exploitation.

  12. Processing of spectral X-ray data with principal components analysis

    CERN Document Server

    Butler, A P H; Cook, N J; Butzer, J; Schleich, N; Tlustos, L; Scott, N; Grasset, R; de Ruiter, N; Anderson, N G

    2011-01-01

    The goal of the work was to develop a general method for processing spectral x-ray image data. Principle component analysis (PCA) is a well understood technique for multivariate data analysis and so was investigated. To assess this method, spectral (multi-energy) computed tomography (CT) data was obtained using a Medipix2 detector in a MARS-CT (Medipix All Resolution System). PCA was able to separate bone (calcium) from two elements with k-edges in the X-ray spectrum used (iodine and barium) within a mouse. This has potential clinical application in dual-energy CT systems and future Medipix3 based spectral imaging where up to eight energies can be recorded simultaneously with excellent energy resolution. (c) 2010 Elsevier B.V. All rights reserved.

  13. Computed tomography of x-ray index of refraction using the diffraction enhanced imaging method

    International Nuclear Information System (INIS)

    Dilmanian, F.A.; Ren, B.; Wu, X.Y.; Orion, I.; Zhong, Z.; Thomlinson, W.C.; Chapman, L.D.

    2000-01-01

    Diffraction enhanced imaging (DEI) is a new, synchrotron-based, x-ray radiography method that uses monochromatic, fan-shaped beams, with an analyser crystal positioned between the subject and the detector. The analyser allows the detection of only those x-rays transmitted by the subject that fall into the acceptance angle (central part of the rocking curve) of the monochromator/analyser system. As shown by Chapman et al , in addition to the x-ray attenuation, the method provides information on the out-of-plane angular deviation of x-rays. New images result in which the image contrast depends on the x-ray index of refraction and on the yield of small-angle scattering, respectively. We implemented DEI in the tomography mode at the National Synchrotron Light Source using 22 keV x-rays, and imaged a cylindrical acrylic phantom that included oil-filled, slanted channels. The resulting 'refraction CT image' shows the pure image of the out-of-plane gradient of the x-ray index of refraction. No image artefacts were present, indicating that the CT projection data were a consistent set. The 'refraction CT image' signal is linear with the gradient of the refractive index, and its value is equal to that expected. The method, at the energy used or higher, has the potential for use in clinical radiography and in industry. (author)

  14. TU-EF-207-01: Introductory Remarks on Recent Advances in Breast Imaging

    Energy Technology Data Exchange (ETDEWEB)

    Karellas, A. [University of Massachusetts Medical School (United States)

    2015-06-15

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation.

  15. TU-EF-207-01: Introductory Remarks on Recent Advances in Breast Imaging

    International Nuclear Information System (INIS)

    Karellas, A.

    2015-01-01

    mode due to lower photon fluence per projection. This may require fast-frame acquisition and symmetric or asymmetric pixel binning in some systems. Recent studies investigated the performance of increased conversion layer thickness for contrast-enhanced imaging of the breast in dual-energy acquisition mode. In other direct conversion detectors operating in the avalanche mode, sensitivities close to the single photon response are also explored for mammography and breast tomosynthesis. The potential advantages and challenges of this approach are described. Dedicated breast CT brings x-ray imaging of the breast to true tomographic 3D imaging. It can eliminate the tissue superposition problem and does not require physical compression of the breast. Using cone beam geometry and a flat-panel detector, several hundred projections are acquired and reconstructed to near isotropic voxels. Multiplanar reconstruction facilitates viewing the breast volume in any desired orientation. Ongoing clinical studies, the current state-of-the art, and research to advance the technology are described. Learning Objectives: To understand the ongoing developments in x-ray imaging of the breast To understand the approaches and applications of spectral mammography To understand the potential advantages of distributed x-ray source arrays for digital breast tomosynthesis To understand the ongoing developments in detector technology for digital mammography and breast tomosynthesis To understand the current state-of-the-art for dedicated cone-beam breast CT and research to advance the technology. Research collaboration with Koning Corporation

  16. The effect of skin thickness determined using breast CT on mammographic dosimetry

    International Nuclear Information System (INIS)

    Huang Shihying; Boone, John M.; Yang, Kai; Kwan, Alexander L. C.; Packard, Nathan J.

    2008-01-01

    The effect of breast skin thickness on dosimetry in mammography was investigated. Breast computed tomography (CT) acquisition techniques, combined with algorithms designed for determining specific breast metrics, were useful for estimating skin thickness. A radial-geometry edge detection scheme was implemented on coronal reconstructed breast CT (bCT) images to measure the breast skin thickness. Skin thickness of bilateral bCT volume data from 49 women and unilateral bCT volume data from 2 women (10 healthy women and 41 women with BIRADS 4 and 5 diagnoses) was robustly measured with the edge detection scheme. The mean breast skin thickness (±inter-breast standard deviation) was found to be 1.45±0.30 mm. Since most current published normalized glandular dose (D gN ) coefficients are based on the assumption of a 4-mm breast skin thickness, the D gN values computed with Monte Carlo techniques will increase up to 18% due to the thinner skin layers (e.g., 6-cm 50% glandular breast, 28 kVp Mo-Mo spectrum). The thinner skin dimensions found in this study suggest that the current D gN values used for mammographic dosimetry lead to a slight underestimate in glandular dose

  17. The "Air in the CT X-ray Tube Oil" Artifact-Examples of the Quality Control Images and the Evaluation of Four Potential Clinical Patients' Head Computed Tomography Cases.

    Science.gov (United States)

    Törmänen, Juhani; Rautiainen, Jari; Tahvonen, Pirita; Leinonen, Kimmo; Nieminen, Miika T; Tervonen, Osmo

    We present a newly reposted scanner-based artifact-with 4 potential patients' head computed tomography (CT) cases-the "Air in the CT X-ray Tube Oil" artifact with a 64-slice multidetector CT. This artifact mimics diseases, which cause hypodense findings in CT images. It can be difficult to notice in the clinical patient imaging but can be also very difficult to verify in quality control tests.

  18. Multiple X-ray tomography using transmitted, scattered and fluorescent radiation

    International Nuclear Information System (INIS)

    Cesareo, R.; Brunetti, A.; Golosio, B.; Lopes, R.T.; Barroso, R.C.; Donativi, M.; Castellano, A.; Quarta, S.

    2003-01-01

    A multiple CT-scanner is described, which contemporaneously uses transmitted, scattered and fluorescent X-rays for Imaging. The scanner is characterized by a small size X-ray tube and by four detectors: a ''pencil'' X-ray NaI(Tl) for transmitted tomography, a larger size NaI(Tl) for 90 C o Compton tomography, a thermoelectrically cooled Si-PIN or CdZnTe for fluorescent imaging and a CdZnTe for Rayleigh (or diffraction) tomography. Examples of applications are shown

  19. Validation of a method for measuring the volumetric breast density from digital mammograms

    International Nuclear Information System (INIS)

    Alonzo-Proulx, O; Shen, S Z; Yaffe, M J; Packard, N; Boone, J M; Al-Mayah, A; Brock, K K

    2010-01-01

    The purpose of this study was to evaluate the performance of an algorithm used to measure the volumetric breast density (VBD) from digital mammograms. The algorithm is based on the calibration of the detector signal versus the thickness and composition of breast-equivalent phantoms. The baseline error in the density from the algorithm was found to be 1.25 ± 2.3% VBD units (PVBD) when tested against a set of calibration phantoms, of thicknesses 3-8 cm, with compositions equivalent to fibroglandular content (breast density) between 0% and 100% and under x-ray beams between 26 kVp and 32 kVp with a Rh/Rh anode/filter. The algorithm was also tested against images from a dedicated breast computed tomography (CT) scanner acquired on 26 volunteers. The CT images were segmented into regions representing adipose, fibroglandular and skin tissues, and then deformed using a finite-element algorithm to simulate the effects of compression in mammography. The mean volume, VBD and thickness of the compressed breast for these deformed images were respectively 558 cm 3 , 23.6% and 62 mm. The displaced CT images were then used to generate simulated digital mammograms, considering the effects of the polychromatic x-ray spectrum, the primary and scattered energy transmitted through the breast, the anti-scatter grid and the detector efficiency. The simulated mammograms were analyzed with the VBD algorithm and compared with the deformed CT volumes. With the Rh/Rh anode filter, the root mean square difference between the VBD from CT and from the algorithm was 2.6 PVBD, and a linear regression between the two gave a slope of 0.992 with an intercept of -1.4 PVBD and a correlation with R 2 = 0.963. The results with the Mo/Mo and Mo/Rh anode/filter were similar.

  20. Validation of a method for measuring the volumetric breast density from digital mammograms

    Energy Technology Data Exchange (ETDEWEB)

    Alonzo-Proulx, O; Shen, S Z; Yaffe, M J [Sunnybrook Health Sciences Centre, University of Toronto, Toronto, Ontario M4N 3M5 (Canada); Packard, N; Boone, J M [UC Davis Medical Center, University of California-Davis, Sacramento, CA 95817 (United States); Al-Mayah, A; Brock, K K, E-mail: oliviera@sri.utoronto.c [University Health Network, University of Toronto, Toronto, Ontario M5G 2M9 (Canada)

    2010-06-07

    The purpose of this study was to evaluate the performance of an algorithm used to measure the volumetric breast density (VBD) from digital mammograms. The algorithm is based on the calibration of the detector signal versus the thickness and composition of breast-equivalent phantoms. The baseline error in the density from the algorithm was found to be 1.25 {+-} 2.3% VBD units (PVBD) when tested against a set of calibration phantoms, of thicknesses 3-8 cm, with compositions equivalent to fibroglandular content (breast density) between 0% and 100% and under x-ray beams between 26 kVp and 32 kVp with a Rh/Rh anode/filter. The algorithm was also tested against images from a dedicated breast computed tomography (CT) scanner acquired on 26 volunteers. The CT images were segmented into regions representing adipose, fibroglandular and skin tissues, and then deformed using a finite-element algorithm to simulate the effects of compression in mammography. The mean volume, VBD and thickness of the compressed breast for these deformed images were respectively 558 cm{sup 3}, 23.6% and 62 mm. The displaced CT images were then used to generate simulated digital mammograms, considering the effects of the polychromatic x-ray spectrum, the primary and scattered energy transmitted through the breast, the anti-scatter grid and the detector efficiency. The simulated mammograms were analyzed with the VBD algorithm and compared with the deformed CT volumes. With the Rh/Rh anode filter, the root mean square difference between the VBD from CT and from the algorithm was 2.6 PVBD, and a linear regression between the two gave a slope of 0.992 with an intercept of -1.4 PVBD and a correlation with R{sup 2} = 0.963. The results with the Mo/Mo and Mo/Rh anode/filter were similar.

  1. Three dimensional dose distribution comparison of simple and complex acquisition trajectories in dedicated breast CT

    Energy Technology Data Exchange (ETDEWEB)

    Shah, Jainil P., E-mail: jainil.shah@duke.edu [Department of Biomedical Engineering, Duke University, Durham, North Carolina 27705 and Multi Modality Imaging Lab, Duke University Medical Center, Durham, North Carolina 27710 (United States); Mann, Steve D. [Medical Physics Graduate Program, Duke University Medical Center, Durham, North Carolina 27705 and Multi Modality Imaging Lab, Duke University Medical Center, Durham, North Carolina 27710 (United States); McKinley, Randolph L. [ZumaTek, Inc., Research Triangle Park, North Carolina 27709 (United States); Tornai, Martin P. [Department of Biomedical Engineering, Duke University, Durham, North Carolina 27705 (United States); Medical Physics Graduate Program, Duke University Medical Center, Durham, North Carolina 27705 (United States); Multi Modality Imaging Lab, Duke University Medical Center, Durham, North Carolina 27710 (United States)

    2015-08-15

    Purpose: A novel breast CT system capable of arbitrary 3D trajectories has been developed to address cone beam sampling insufficiency as well as to image further into the patient’s chest wall. The purpose of this study was to characterize any trajectory-related differences in 3D x-ray dose distribution in a pendant target when imaged with different orbits. Methods: Two acquisition trajectories were evaluated: circular azimuthal (no-tilt) and sinusoidal (saddle) orbit with ±15° tilts around a pendant breast, using Monte Carlo simulations as well as physical measurements. Simulations were performed with tungsten (W) filtration of a W-anode source; the simulated source flux was normalized to the measured exposure of a W-anode source. A water-filled cylindrical phantom was divided into 1 cm{sup 3} voxels, and the cumulative energy deposited was tracked in each voxel. Energy deposited per voxel was converted to dose, yielding the 3D distributed dose volumes. Additionally, three cylindrical phantoms of different diameters (10, 12.5, and 15 cm) and an anthropomorphic breast phantom, initially filled with water (mimicking pure fibroglandular tissue) and then with a 75% methanol-25% water mixture (mimicking 50–50 fibroglandular-adipose tissues), were used to simulate the pendant breast geometry and scanned on the physical system. Ionization chamber calibrated radiochromic film was used to determine the dose delivered in a 2D plane through the center of the volume for a fully 3D CT scan using the different orbits. Results: Measured experimental results for the same exposure indicated that the mean dose measured throughout the central slice for different diameters ranged from 3.93 to 5.28 mGy, with the lowest average dose measured on the largest cylinder with water mimicking a homogeneously fibroglandular breast. These results align well with the cylinder phantom Monte Carlo studies which also showed a marginal difference in dose delivered by a saddle trajectory in the

  2. Three dimensional dose distribution comparison of simple and complex acquisition trajectories in dedicated breast CT

    International Nuclear Information System (INIS)

    Shah, Jainil P.; Mann, Steve D.; McKinley, Randolph L.; Tornai, Martin P.

    2015-01-01

    Purpose: A novel breast CT system capable of arbitrary 3D trajectories has been developed to address cone beam sampling insufficiency as well as to image further into the patient’s chest wall. The purpose of this study was to characterize any trajectory-related differences in 3D x-ray dose distribution in a pendant target when imaged with different orbits. Methods: Two acquisition trajectories were evaluated: circular azimuthal (no-tilt) and sinusoidal (saddle) orbit with ±15° tilts around a pendant breast, using Monte Carlo simulations as well as physical measurements. Simulations were performed with tungsten (W) filtration of a W-anode source; the simulated source flux was normalized to the measured exposure of a W-anode source. A water-filled cylindrical phantom was divided into 1 cm"3 voxels, and the cumulative energy deposited was tracked in each voxel. Energy deposited per voxel was converted to dose, yielding the 3D distributed dose volumes. Additionally, three cylindrical phantoms of different diameters (10, 12.5, and 15 cm) and an anthropomorphic breast phantom, initially filled with water (mimicking pure fibroglandular tissue) and then with a 75% methanol-25% water mixture (mimicking 50–50 fibroglandular-adipose tissues), were used to simulate the pendant breast geometry and scanned on the physical system. Ionization chamber calibrated radiochromic film was used to determine the dose delivered in a 2D plane through the center of the volume for a fully 3D CT scan using the different orbits. Results: Measured experimental results for the same exposure indicated that the mean dose measured throughout the central slice for different diameters ranged from 3.93 to 5.28 mGy, with the lowest average dose measured on the largest cylinder with water mimicking a homogeneously fibroglandular breast. These results align well with the cylinder phantom Monte Carlo studies which also showed a marginal difference in dose delivered by a saddle trajectory in the

  3. Clinical implementation of a low energy x-ray therapy device in the treatment of breast cancer

    International Nuclear Information System (INIS)

    Haworth, A.; University of Western Australia, WA; Joseph, D.; Lanzon, P.; Caswell, N.; Ebert, M.; University of Western Asutralia, WA

    2001-01-01

    Full text: A low energy device producing x-rays of maximum operating potential of 50kV is used to treat primary breast tumours intraoperatively. In pathologically favourable cases, the treatment replaces conventional external beam irradiation. For patients at greater risk of local recurrence, the treatment replaces conventional 'boost' therapy. The dosimetry of the device will be described in a companion paper. QA tests prior to irradiation include: output calibration/verification; isotropy verification and external radiation monitor (the secondary beam termination device) functionality. The internal radiation monitor count (similar to setting monitor units on a linac) for a prescribed dose is calculated from tables of measured depth dose and applicator factors. The spherical applicator which best suits the size of the excised tumour is lightly sutured into position maintaining as much distance between the skin surface as possible to minimise skin erythema. Radiation protection is achieved with the use of portable lead shields and tungsten impregnated silicon drapes. Patients entered into a TROG randomised clinical trial comparing intraoperative with conventional postoperative radiotherapy after conservative breast surgery for women with early stage breast cancer will be studied in collaboration with the CRC/University College London, Cancer Trials Centre (UK) to record the effects of local tumour control, cosmesis, patient satisfaction and health economics. QA tests take approximately 15 minutes to perform and a treatment prescription of 5Gy at 1cm depth with a 5cm applicator would take approximately 30 minutes. A low energy x-ray device may be used intraoperatively in selected cases to replace conventional radiotherapy minimising the inconvenience for patients and reducing waiting lists on treatment machines. Copyright (2001) Australasian College of Physical Scientists and Engineers in Medicine

  4. A minimum spanning forest based classification method for dedicated breast CT images

    International Nuclear Information System (INIS)

    Pike, Robert; Sechopoulos, Ioannis; Fei, Baowei

    2015-01-01

    Purpose: To develop and test an automated algorithm to classify different types of tissue in dedicated breast CT images. Methods: Images of a single breast of five different patients were acquired with a dedicated breast CT clinical prototype. The breast CT images were processed by a multiscale bilateral filter to reduce noise while keeping edge information and were corrected to overcome cupping artifacts. As skin and glandular tissue have similar CT values on breast CT images, morphologic processing is used to identify the skin based on its position information. A support vector machine (SVM) is trained and the resulting model used to create a pixelwise classification map of fat and glandular tissue. By combining the results of the skin mask with the SVM results, the breast tissue is classified as skin, fat, and glandular tissue. This map is then used to identify markers for a minimum spanning forest that is grown to segment the image using spatial and intensity information. To evaluate the authors’ classification method, they use DICE overlap ratios to compare the results of the automated classification to those obtained by manual segmentation on five patient images. Results: Comparison between the automatic and the manual segmentation shows that the minimum spanning forest based classification method was able to successfully classify dedicated breast CT image with average DICE ratios of 96.9%, 89.8%, and 89.5% for fat, glandular, and skin tissue, respectively. Conclusions: A 2D minimum spanning forest based classification method was proposed and evaluated for classifying the fat, skin, and glandular tissue in dedicated breast CT images. The classification method can be used for dense breast tissue quantification, radiation dose assessment, and other applications in breast imaging

  5. Trends in x-ray photography and patient exposure dose

    International Nuclear Information System (INIS)

    Orito, Takeo; Sanada, Shigeru; Maekawa, Ryuichi; Koshida, Kichiro; Hiraki, Tatsunosuke

    1980-01-01

    The exposure doses of patients in X-ray photography are influenced by such technological factors as X-ray tube voltage, filter, sensitizing screen, film and grid. Survey by questionnnaire was made previously in 1973 on the above factors. The trends five years after were surveyed similarly, in connection with the exposure doses of patients. Questionnaires were sent to 200 radiation technicians, and 121 (60.5%) answered the survey in March, 1979. The results in the cases of simple X-ray photography and obstetric, infant and breast X-ray photographings are described. X-ray tube voltage is generally on the increase. In the sensitizing screens, exposure doses are fairly decreased due to the use of improved intensifying screen (LT-II). In the grid, the ratio 8 : 1 is used more than 5 : 1. In the usage of additional filters and in the distance of photography, improvements are desired. (J.P.N.)

  6. Diagnostic perspectives of mobile x-ray computerized tomography

    International Nuclear Information System (INIS)

    Portnoj, L.M.; Dibirov, M.P.; Denisova, L.B.

    1992-01-01

    Mobile x-ray computerized tomography (CT) and an organization and methodological scheme of its application were assessed. CT is realized in special hospitals of large regions, where the patients with the optimal indications for CT are assembled. Over 15000 examinations were carried out with the use of the suggested CT program over 4 years, that resulted in detection of 1295 brain tumors, 804 cases with neoplastic involvement of the abdominal cavity and the retroperitoneal space. Wide application of mobile CT devices according to the program will help to decide the problem of unavailability of such examinations. One mobile device may replace 3 permanent CT devices

  7. Emphysema quantification and lung volumetry in chest X-ray equivalent ultralow dose CT - Intra-individual comparison with standard dose CT.

    Science.gov (United States)

    Messerli, Michael; Ottilinger, Thorsten; Warschkow, René; Leschka, Sebastian; Alkadhi, Hatem; Wildermuth, Simon; Bauer, Ralf W

    2017-06-01

    To determine whether ultralow dose chest CT with tin filtration can be used for emphysema quantification and lung volumetry and to assess differences in emphysema measurements and lung volume between standard dose and ultralow dose CT scans using advanced modeled iterative reconstruction (ADMIRE). 84 consecutive patients from a prospective, IRB-approved single-center study were included and underwent clinically indicated standard dose chest CT (1.7±0.6mSv) and additional single-energy ultralow dose CT (0.14±0.01mSv) at 100kV and fixed tube current at 70mAs with tin filtration in the same session. Forty of the 84 patients (48%) had no emphysema, 44 (52%) had emphysema. One radiologist performed fully automated software-based pulmonary emphysema quantification and lung volumetry of standard and ultralow dose CT with different levels of ADMIRE. Friedman test and Wilcoxon rank sum test were used for multiple comparison of emphysema and lung volume. Lung volumes were compared using the concordance correlation coefficient. The median low-attenuation areas (LAA) using filtered back projection (FBP) in standard dose was 4.4% and decreased to 2.6%, 2.1% and 1.8% using ADMIRE 3, 4, and 5, respectively. The median values of LAA in ultralow dose CT were 5.7%, 4.1% and 2.4% for ADMIRE 3, 4, and 5, respectively. There was no statistically significant difference between LAA in standard dose CT using FBP and ultralow dose using ADMIRE 4 (p=0.358) as well as in standard dose CT using ADMIRE 3 and ultralow dose using ADMIRE 5 (p=0.966). In comparison with standard dose FBP the concordance correlation coefficients of lung volumetry were 1.000, 0.999, and 0.999 for ADMIRE 3, 4, and 5 in standard dose, and 0.972 for ADMIRE 3, 4 and 5 in ultralow dose CT. Ultralow dose CT at chest X-ray equivalent dose levels allows for lung volumetry as well as detection and quantification of emphysema. However, longitudinal emphysema analyses should be performed with the same scan protocol and

  8. Development of ''Eminence STARGATE'' PET/CT system

    International Nuclear Information System (INIS)

    Okazaki, Masato; Inoue, Yoshihiro; Amano, Masaharu

    2009-01-01

    A PET/CT system, the combination of a PET (Positron Emission Tomography) system with an X-ray CT system, has been widely used in recent years. Our newly developed ''Eminence STARGATE'' PET/CT system allows the PET gantry and the X-ray CT gantry to move independently. This advantage provides high flexibility for PET examination and X-ray CT examination and also eases a patient's psychological anxiety about closed spaces. The system has a 16-slice X-ray CT scanner. (author)

  9. Assessment of fluid distribution and flow properties in two phase fluid flow using X-ray CT technology

    Science.gov (United States)

    Jiang, Lanlan; Wu, Bohao; Li, Xingbo; Wang, Sijia; Wang, Dayong; Zhou, Xinhuan; Zhang, Yi

    2018-04-01

    To study on microscale distribution of CO2 and brine during two-phase flow is crucial for understanding the trapping mechanisms of CO2 storage. In this study, CO2-brine flow experiments in porous media were conducted using X-ray computed tomography. The porous media were packed with glass beads. The pore structure (porosity/tortuosity) and flow properties at different flow rates and flow fractions were investigated. The results showed that porosity of the packed beads differed at different position as a result of heterogeneity. The CO2 saturation is higher at low injection flow rates and high CO2 fractions. CO2 distribution at the pore scale was also visualized. ∅ Porosity of porous media CT brine_ sat grey value of sample saturated with brine CT dry grey value of sample saturated with air CT brine grey value of pure brine CT air grey value of pure air CT flow grey values of sample with two fluids occupying the pore space {CT}_{CO_2_ sat} grey value of sample saturated with CO2 {f}_{CO_2}({S}_{CO_2}) CO2 fraction {q}_{CO_2} the volume flow rate for CO2 q brine the volume flow rate for brine L Thickness of the porous media, mm L e a bundle of capillaries of equal length, mm τ Tortuosity, calculated from L e / L.

  10. Evaluation of accelerated iterative x-ray CT image reconstruction using floating point graphics hardware

    International Nuclear Information System (INIS)

    Kole, J S; Beekman, F J

    2006-01-01

    Statistical reconstruction methods offer possibilities to improve image quality as compared with analytical methods, but current reconstruction times prohibit routine application in clinical and micro-CT. In particular, for cone-beam x-ray CT, the use of graphics hardware has been proposed to accelerate the forward and back-projection operations, in order to reduce reconstruction times. In the past, wide application of this texture hardware mapping approach was hampered owing to limited intrinsic accuracy. Recently, however, floating point precision has become available in the latest generation commodity graphics cards. In this paper, we utilize this feature to construct a graphics hardware accelerated version of the ordered subset convex reconstruction algorithm. The aims of this paper are (i) to study the impact of using graphics hardware acceleration for statistical reconstruction on the reconstructed image accuracy and (ii) to measure the speed increase one can obtain by using graphics hardware acceleration. We compare the unaccelerated algorithm with the graphics hardware accelerated version, and for the latter we consider two different interpolation techniques. A simulation study of a micro-CT scanner with a mathematical phantom shows that at almost preserved reconstructed image accuracy, speed-ups of a factor 40 to 222 can be achieved, compared with the unaccelerated algorithm, and depending on the phantom and detector sizes. Reconstruction from physical phantom data reconfirms the usability of the accelerated algorithm for practical cases

  11. Morphological and quantitative analyses on features of asymmetric mandible with X-ray CT images

    Energy Technology Data Exchange (ETDEWEB)

    Hayashi, Masaki; Isshiki, Yasushige; Nishikawa, Keiichi [Tokyo Dental Coll., Chiba (Japan)

    2001-09-01

    Biomechanical disproportion of the mandible causes morphological and quantitative changes in the bone and induces jaw deformities. The purpose of this study was to clarify the influence of the biomechanical disproportion on the structure and quality of the mandible, using spiral X-ray CT images. Subjects were 11 patients with mandibular asymmetry requiring orthognathic surgery. Lateral and vertical shift angles of the mandible against the maxilla were determined mathematically from the coordinates of structural reference points in 3-dimensional orthogonal coordinate systems as indices of the deformity degree. Several properties concerning thickness and CT value of mandibular cortical bone were measured as indices of the morphological and quantitative changes in mandibular cortical bone. Occlusal force was also measured using a pressure sensitive film as an index of biomechanical disproportion. Asymmetric indices of them were calculated from data of the left and right sides. The Pearson's correlation coefficients were obtained for these data. As a result, the mandible tended to shift laterally toward the side where the occlusal force at the second molar region was stronger and also to shift upward on that side. At the central incisor region of the laterally shifted side, CT values were relatively decreased. At the lateral incisor region, cortical bone density was relatively increased. At the first bicuspid region and the posterior tooth regions, cortical bone thickness was relatively decreased. (author)

  12. Morphological and quantitative analyses on features of asymmetric mandible with X-ray CT images

    International Nuclear Information System (INIS)

    Hayashi, Masaki; Isshiki, Yasushige; Nishikawa, Keiichi

    2001-01-01

    Biomechanical disproportion of the mandible causes morphological and quantitative changes in the bone and induces jaw deformities. The purpose of this study was to clarify the influence of the biomechanical disproportion on the structure and quality of the mandible, using spiral X-ray CT images. Subjects were 11 patients with mandibular asymmetry requiring orthognathic surgery. Lateral and vertical shift angles of the mandible against the maxilla were determined mathematically from the coordinates of structural reference points in 3-dimensional orthogonal coordinate systems as indices of the deformity degree. Several properties concerning thickness and CT value of mandibular cortical bone were measured as indices of the morphological and quantitative changes in mandibular cortical bone. Occlusal force was also measured using a pressure sensitive film as an index of biomechanical disproportion. Asymmetric indices of them were calculated from data of the left and right sides. The Pearson's correlation coefficients were obtained for these data. As a result, the mandible tended to shift laterally toward the side where the occlusal force at the second molar region was stronger and also to shift upward on that side. At the central incisor region of the laterally shifted side, CT values were relatively decreased. At the lateral incisor region, cortical bone density was relatively increased. At the first bicuspid region and the posterior tooth regions, cortical bone thickness was relatively decreased. (author)

  13. Intraoperative Radiation Using Low-Kilovoltage X-Rays for Early Breast Cancer: A Single Site Trial.

    Science.gov (United States)

    Silverstein, Melvin J; Epstein, Melinda S; Lin, Kevin; Chen, Peter; Khan, Sadia; Snyder, Lincoln; Guerra, Lisa E; De Leon, Cristina; Mackintosh, Ralph; Coleman, Colleen; Lopez, January; Kim, Brian

    2017-10-01

    Two prospective, randomized trials, TARGIT-A and ELIOT, have shown intraoperative radiation therapy (IORT) to be a safe alternative to whole breast radiation therapy following breast-conserving surgery for selected low-risk patients. However, minimal data are available about the clinical effectiveness of this modality of treatment using the Xoft ® Axxent ® Electronic Brachytherapy (eBx ® ) System ® . A total of 201 patients with 204 early-stage breast cancers were enrolled in a prospective X-ray IORT trial from June 2010 to September 2013. All tumors were treated with breast-conserving surgery and IORT. Data were collected at 1 week, 1 month, 6 months, 1 year, and yearly thereafter. With a median follow-up of 50 months, there have been seven ipsilateral breast tumor events (IBTE), no regional or distant recurrences, and no breast cancer-related deaths. One IBTE was within the IORT field, four outside of the IORT field but within the same quadrant as the index cancer, and two were new biologically different cancers in different quadrants. Three events were in patients who deviated from the protocol criteria. Kaplan-Meier analysis projects that 2.9% of patients will recur locally at 4 years. Recurrence rates observed in this trial were comparable to those of the TARGIT-A and ELIOT trials as well as the retrospective TARGIT-R trial. The low complication rates previously reported by our group as well as the low recurrence rates reported in this study support the cautious use and continued study of IORT in selected women with low-risk breast cancer.

  14. Performance of FDG PET/CT in the clinical management of breast cancer.

    Science.gov (United States)

    Groheux, David; Espié, Marc; Giacchetti, Sylvie; Hindié, Elif

    2013-02-01

    In this analysis, the role of metabolic imaging with fluorine 18 fluorodeoxyglucose (FDG) in breast cancer is reviewed. The analysis was limited to recent works by using state-of-the-art positron emission tomography (PET)/computed tomography (CT) technology. The strengths and limitations of FDG PET/CT are examined in various clinical settings, and the following questions are answered: Is FDG PET/CT useful to differentiate malignant from benign breast lesions? Can FDG PET/CT replace sentinel node biopsy for axillary staging? What is the role of FDG PET/CT in initial staging of inflammatory or locally advanced breast cancer? What is the role of FDG PET/CT in initial staging of clinical stage IIA and IIB and primary operable stage IIIA breast cancer? How does FDG PET/CT compare with conventional techniques in the restaging of cancer in patients who are suspected of having disease recurrence? What is the role of FDG PET/CT in the assessment of early response to neoadjuvant therapy and of response to therapy for metastatic disease? Some recommendations for clinical practice are given.

  15. In vivo tomographic imaging of lung colonization of tumour in mouse with simultaneous fluorescence and X-ray CT.

    Science.gov (United States)

    Zhang, Bin; Gao, Fuping; Wang, Mengjiao; Cao, Xu; Liu, Fei; Wang, Xin; Luo, Jianwen; Wang, Guangzhi; Bai, Jing

    2014-01-01

    Non-invasive in vivo imaging of diffuse and wide-spread colonization within the lungs, rather than distinct solid primary tumors, is still a challenging work. In this work, a lung colonization mouse model bearing A549 human lung tumor was simultaneously scanned by a dual-modality fluorescence molecular tomography (FMT) and X-ray computed tomography (CT) system in vivo. A two steps method which incorporates CT structural information into the FMT reconstruction procedure is employed to provide concurrent anatomical and functional information. By using the target-specific fluorescence agent, the fluorescence tomographic results show elevated fluorescence intensity deep within the lungs which is colonized with diffuse and wide-spread tumors. The results were confirmed with ex vivo fluorescence reflectance imaging and histological examination of the lung tissues. With FMT reconstruction combined with the CT information, the dual-modality FMT/micro-CT system is expected to offer sensitive and noninvasive imaging of diffuse tumor colonization within the lungs in vivo. Copyright © 2014 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim.

  16. Anatomical accuracy of lesion localization. Retrospective interactive rigid image registration between 18F-FDG-PET and X-ray CT

    International Nuclear Information System (INIS)

    Noemayr, A.; Roemer, W.; Kuwert, T.; Hothorn, T.; Pfahlberg, A.; Hornegger, J.; Bautz, W.

    2005-01-01

    The aim of this study was to evaluate the anatomical accuracy and reproducibility of retrospective interactive rigid image registration (RIR) between routinely archived X-ray computer tomography (CT) and positron emission tomography performed with 18 F-deoxyglucose (FDG-PET) in oncological patients. Methods: two observers registered PET and CT data obtained in 37 patients using a commercially available image fusion tool. RIR was performed separately for the thorax and the abdomen using physiological FDG uptake in several organs as a reference. One observer performed the procedure twice (01a and 01b), another person once (02). For 94 malignant lesions, clearly visible in CT and PET, the signed and absolute distances between their representation on PET and CT were measured in X-, Y-, and Z-direction with reference to a coordinate system centered in the CT representation of each lesion (X-, Y-, Z-distances). Results: the mean differences of the signed and absolute distances between 01a, 01b, and 02 did not exceed 3 mm in any dimension. The absolute X-, Y-, and Z-distances ranged between 0.57 ± 0.58 cm for 01a (X-direction) and 1.12 ± 1.28 cm for 02 (Z-direction). When averaging the absolute distances measured by 01a, 01b, and 02, the percentage of lesions misregistered by less than 1.5 cm was 91% for the X-, 88% for the Y-, and 77% for the Z-direction. The larger error of fusion determined for the remaining lesions was caused by non-rigid body transformations due to differences in breathing, arm position, or bowel movements between the two examinations. Mixed effects analysis of the signed and absolute X-, Y-, and Z-distances disclosed a significantly greater misalignment in the thorax than in the abdomen as well as axially than transaxially. Conclusion: the anatomical inaccuracy of RIR can be expected to be <1.5 cm for the majority of neoplastic foci. errors of alignment are bigger in the thorax and in Z-direction, due to non-rigid body transformations caused, e

  17. Molecular Imaging in Breast Cancer: From Whole-Body PET/CT to Dedicated Breast PET

    Directory of Open Access Journals (Sweden)

    B. B. Koolen

    2012-01-01

    Full Text Available Positron emission tomography (PET, with or without integrated computed tomography (CT, using 18F-fluorodeoxyglucose (FDG is based on the principle of elevated glucose metabolism in malignant tumors, and its use in breast cancer patients is frequently being investigated. It has been shown useful for classification, staging, and response monitoring, both in primary and recurrent disease. However, because of the partial volume effect and limited resolution of most whole-body PET scanners, sensitivity for the visualization of small tumors is generally low. To improve the detection and quantification of primary breast tumors with FDG PET, several dedicated breast PET devices have been developed. In this nonsystematic review, we shortly summarize the value of whole-body PET/CT in breast cancer and provide an overview of currently available dedicated breast PETs.

  18. Quantitative Study of Porosity and Pore Features in Moldavites by Means of X-ray Micro-CT

    Directory of Open Access Journals (Sweden)

    Giovanni Pratesi

    2014-04-01

    Full Text Available X-ray micro-computer aided tomography (μ-CT, together with optical microscopy and imaging, have been applied to the study of six moldavite samples. These techniques enabled a complete characterization to be made of the textural features of both Muong Nong-type and common splashform moldavites. A detailed study of the size and distribution of pores or bubbles confirmed the marked variability in pore size among the samples, as well as within each sample, and indicated in the Muong Nong-type moldavites the presence of at least two deformation stages which occurred before and after pore formation.

  19. Eighteen cases of small breast cancer: a comparative study of mammography, CT scan and pathology

    International Nuclear Information System (INIS)

    Wu Yaopan; Lin Haogao; Cai Peiqiang; Ouyang Yi; Zhang Weizhang; Lu Bingui

    2003-01-01

    Objective: To improve the early diagnosis of breast cancer through a study of the mammography and CT findings of small breast cancer. Methods: The mammography and CT findings of 18 cases of small breast cancer (φ≤2.0 cm in diameter) were studied and compared with pathological results. Results: The diagnostic accuracy of CT and mammography was 83% and 61%, respectively. There was a statistical difference between both modalities (P<0.05), CT scan was superior to mammography. However, there was no difference between them when assessing the lesion arising in F-type breast. In detecting breast fine cluster of calcification, the sensitivity of mammography was better than CT scan. Conclusion: The patient suspected of small breast cancer should take mammography as the first evaluation. CT scan is reserved for the further investigation. The mammography combined with CT scan can improve the early diagnostic rate of breast cancer

  20. Characterization of metal additive manufacturing surfaces using synchrotron X-ray CT and micromechanical modeling

    Science.gov (United States)

    Kantzos, C. A.; Cunningham, R. W.; Tari, V.; Rollett, A. D.

    2018-05-01

    Characterizing complex surface topologies is necessary to understand stress concentrations created by rough surfaces, particularly those made via laser power-bed additive manufacturing (AM). Synchrotron-based X-ray microtomography (μ XCT) of AM surfaces was shown to provide high resolution detail of surface features and near-surface porosity. Using the CT reconstructions to instantiate a micromechanical model indicated that surface notches and near-surface porosity both act as stress concentrators, while adhered powder carried little to no load. Differences in powder size distribution had no direct effect on the relevant surface features, nor on stress concentrations. Conventional measurements of surface roughness, which are highly influenced by adhered powder, are therefore unlikely to contain the information relevant to damage accumulation and crack initiation.

  1. X-ray detectors in medical imaging

    International Nuclear Information System (INIS)

    Spahn, Martin

    2013-01-01

    Healthcare systems are subject to continuous adaptation, following trends such as the change of demographic structures, the rise of life-style related and chronic diseases, and the need for efficient and outcome-oriented procedures. This also influences the design of new imaging systems as well as their components. The applications of X-ray imaging in the medical field are manifold and have led to dedicated modalities supporting specific imaging requirements, for example in computed tomography (CT), radiography, angiography, surgery or mammography, delivering projection or volumetric imaging data. Depending on the clinical needs, some X-ray systems enable diagnostic imaging while others support interventional procedures. X-ray detector design requirements for the different medical applications can vary strongly with respect to size and shape, spatial resolution, frame rates and X-ray flux, among others. Today, integrating X-ray detectors are in common use. They are predominantly based on scintillators (e.g. CsI or Gd 2 O 2 S) and arrays of photodiodes made from crystalline silicon (Si) or amorphous silicon (a-Si) or they employ semiconductors (e.g. Se) with active a-Si readout matrices. Ongoing and future developments of X-ray detectors will include optimization of current state-of-the-art integrating detectors in terms of performance and cost, will enable the usage of large size CMOS-based detectors, and may facilitate photon counting techniques with the potential to further enhance performance characteristics and foster the prospect of new clinical applications

  2. Preliminary Examination of X-ray Scattering from Human Tissues

    International Nuclear Information System (INIS)

    Desouky, O.S.; Wilkinson, S.; Hall, C.; Rogers, K.; Round, A.

    2008-01-01

    Small Angle x-ray scattering (SAXS) and wide angle x-ray scattering (WAXS) patterns have been recorded from different human soft tissues using x-ray synchrotron radiation.Pathological breast, normal kidney and lung tissues show SAXS peaks at q-values equal to 0.291 nm -1 and 0.481 nm -1 (d 21.6 nm and d =13. nm) which are the 3 r d and 5 t h order of the well known axial D-spacing of collagen fibrils. The diffraction is particularly intense in the meridional direction indicating some febrile alignment. In contrast, the normal tissue of brain, liver and heart shows diffuse scatter.The wide-angle coherent scattering from normal human tissues of brain, liver, heart, lung, and kidney is typical of that for amorphous materials. The scatter of the healthy adipose breast tissue shows a sharp peak at momentum transfer 1.24 nm -1 (d= 0.417 nm). The data of the other tissues appears to consist of a broad scattering peak. The two scattering regimes succeed in differentiating between the two major components of breast tissue, collagen and adipose tissue. The results of this study suggest that the soft tissues may have scattering patterns that are characteristics for the particular tissue types and tissue disease state. These results indicate that it may be possible use the coherent scattering as a diagnostic tool

  3. A semiempirical linear model of indirect, flat-panel x-ray detectors

    Energy Technology Data Exchange (ETDEWEB)

    Huang, Shih-Ying; Yang Kai; Abbey, Craig K.; Boone, John M. [Department of Biomedical Engineering, University of California, Davis, California, One Shields Avenue, Davis, California 95616 (United States) and Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 0505, Sacramento, California 95817 (United States); Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 0505, Sacramento, California 95817 (United States); Department of Psychological and Brain Sciences, University of California, Santa Barbara, California 92106 (United States); Department of Biomedical Engineering, University of California, Davis, California, One Shields Avenue, Davis, California 95616 (United States) and Department of Radiology, University of California, Davis, Medical Center, 4860 Y Street, Ambulatory Care Center Suite 3100, Sacramento, California 95817 (United States)

    2012-04-15

    Purpose: It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. Methods: The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV/0.1-mm Sn, 65 kV/0.2-mm Cu, 85 kV/1.5-mm Al, and 95 kV/0.05-mm Ag. Results: The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r

  4. A semiempirical linear model of indirect, flat-panel x-ray detectors

    International Nuclear Information System (INIS)

    Huang, Shih-Ying; Yang Kai; Abbey, Craig K.; Boone, John M.

    2012-01-01

    Purpose: It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. Methods: The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV/0.1-mm Sn, 65 kV/0.2-mm Cu, 85 kV/1.5-mm Al, and 95 kV/0.05-mm Ag. Results: The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r 2 of

  5. A semiempirical linear model of indirect, flat-panel x-ray detectors.

    Science.gov (United States)

    Huang, Shih-Ying; Yang, Kai; Abbey, Craig K; Boone, John M

    2012-04-01

    It is important to understand signal and noise transfer in the indirect, flat-panel x-ray detector when developing and optimizing imaging systems. For optimization where simulating images is necessary, this study introduces a semiempirical model to simulate projection images with user-defined x-ray fluence interaction. The signal and noise transfer in the indirect, flat-panel x-ray detectors is characterized by statistics consistent with energy-integration of x-ray photons. For an incident x-ray spectrum, x-ray photons are attenuated and absorbed in the x-ray scintillator to produce light photons, which are coupled to photodiodes for signal readout. The signal mean and variance are linearly related to the energy-integrated x-ray spectrum by empirically determined factors. With the known first- and second-order statistics, images can be simulated by incorporating multipixel signal statistics and the modulation transfer function of the imaging system. To estimate the semiempirical input to this model, 500 projection images (using an indirect, flat-panel x-ray detector in the breast CT system) were acquired with 50-100 kilovolt (kV) x-ray spectra filtered with 0.1-mm tin (Sn), 0.2-mm copper (Cu), 1.5-mm aluminum (Al), or 0.05-mm silver (Ag). The signal mean and variance of each detector element and the noise power spectra (NPS) were calculated and incorporated into this model for accuracy. Additionally, the modulation transfer function of the detector system was physically measured and incorporated in the image simulation steps. For validation purposes, simulated and measured projection images of air scans were compared using 40 kV∕0.1-mm Sn, 65 kV∕0.2-mm Cu, 85 kV∕1.5-mm Al, and 95 kV∕0.05-mm Ag. The linear relationship between the measured signal statistics and the energy-integrated x-ray spectrum was confirmed and incorporated into the model. The signal mean and variance factors were linearly related to kV for each filter material (r(2) of signal mean to k

  6. Large area CMOS active pixel sensor x-ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization.

    Science.gov (United States)

    Zhao, Chumin; Kanicki, Jerzy; Konstantinidis, Anastasios C; Patel, Tushita

    2015-11-01

    Large area x-ray imagers based on complementary metal-oxide-semiconductor (CMOS) active pixel sensor (APS) technology have been proposed for various medical imaging applications including digital breast tomosynthesis (DBT). The low electronic noise (50-300 e-) of CMOS APS x-ray imagers provides a possible route to shrink the pixel pitch to smaller than 75 μm for microcalcification detection and possible reduction of the DBT mean glandular dose (MGD). In this study, imaging performance of a large area (29×23 cm2) CMOS APS x-ray imager [Dexela 2923 MAM (PerkinElmer, London)] with a pixel pitch of 75 μm was characterized and modeled. The authors developed a cascaded system model for CMOS APS x-ray imagers using both a broadband x-ray radiation and monochromatic synchrotron radiation. The experimental data including modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE) were theoretically described using the proposed cascaded system model with satisfactory consistency to experimental results. Both high full well and low full well (LFW) modes of the Dexela 2923 MAM CMOS APS x-ray imager were characterized and modeled. The cascaded system analysis results were further used to extract the contrast-to-noise ratio (CNR) for microcalcifications with sizes of 165-400 μm at various MGDs. The impact of electronic noise on CNR was also evaluated. The LFW mode shows better DQE at low air kerma (Ka<10 μGy) and should be used for DBT. At current DBT applications, air kerma (Ka∼10 μGy, broadband radiation of 28 kVp), DQE of more than 0.7 and ∼0.3 was achieved using the LFW mode at spatial frequency of 0.5 line pairs per millimeter (lp/mm) and Nyquist frequency ∼6.7 lp/mm, respectively. It is shown that microcalcifications of 165-400 μm in size can be resolved using a MGD range of 0.3-1 mGy, respectively. In comparison to a General Electric GEN2 prototype DBT system (at MGD of 2.5 mGy), an increased CNR (by ∼10) for

  7. Determination of liquid's molecular interference function based on X-ray diffraction and dual-energy CT in security screening

    International Nuclear Information System (INIS)

    Zhang, Li; YangDai, Tianyi

    2016-01-01

    A method for deriving the molecular interference function (MIF) of an unknown liquid for security screening is presented. Based on the effective atomic number reconstructed from dual-energy computed tomography (CT), equivalent molecular formula of the liquid is estimated. After a series of optimizations, the MIF and a new effective atomic number are finally obtained from the X-ray diffraction (XRD) profile. The proposed method generates more accurate results with less sensitivity to the noise and data deficiency of the XRD profile. - Highlights: • EDXRD combined with dual-energy CT has been utilized for deriving the molecular interference function of an unknown liquid. • The liquid's equivalent molecular formula is estimated based on the effective atomic number reconstructed from dual-energy CT. • The proposed method provides two ways to estimate the molecular interference function: the simplified way and accurate way. • A new effective atomic number of the liquid could be obtained.

  8. Synthesis of PEG-Iodine-Capped Gold Nanoparticles and Their Contrast Enhancement in In Vitro and In Vivo for X-Ray/CT

    International Nuclear Information System (INIS)

    Sun-Hee, K.; Eun-Mi, K.; Chang-Moon, L.; Dong, W.K.; Seok, T.L.; Myung-Hee, S.; Hwan-Jeong, J.

    2012-01-01

    We designed gold nanoparticles (AuNPs) capped with iodine and polyethylene glycol (PEG) to provide effective enhancement for X-ray CT imaging. The methoxy PEG-iodine-capped AuNPs were prepared through the chemisorption of iodine and substitution of methoxy PEG-SH onto the surface of gold nanoparticles, and severe aggregation in TEM was not observed. The binding energies of Au 4f 7/2 and I 3d 5/2 of the methoxy PEG-iodine-capped AuNPs were obtained as 84.1 eV and 619.3 eV, respectively. The binding energy shift of methoxy PEG-iodine-capped AuNPs would be resulted from the chemisorption between gold nanoparticles and iodine atoms. The methoxy PEG-iodine-capped AuNPs have higher enhancement compared to PEG-capped gold nanoparticles in the same amount of gold in vitro. After postinjection of methoxy PEG-iodine-capped AuNPs into the mice, dramatic contrast enhancement at the heart, aorta, liver, and kidney was observed, this was maintained up to 5 days, and there was no evidence of apparent toxicity. In conclusion, methoxy PEG-iodine-capped AuNPs might be a good candidate as a CT contrast agent for blood pool imaging, and this will also contribute to the prolongation of a blood circulation time for X-ray CT imaging.

  9. Real-time fusion of coronary CT angiography with X-ray fluoroscopy during chronic total occlusion PCI

    Energy Technology Data Exchange (ETDEWEB)

    Ghoshhajra, Brian B.; Takx, Richard A.P. [Harvard Medical School, Cardiac MR PET CT Program, Massachusetts General Hospital, Department of Radiology and Division of Cardiology, Boston, MA (United States); Stone, Luke L.; Yeh, Robert W.; Jaffer, Farouc A. [Harvard Medical School, Cardiac Cathetrization Laboratory, Cardiology Division, Massachusetts General Hospital, Boston, MA (United States); Girard, Erin E. [Siemens Healthcare, Princeton, NJ (United States); Brilakis, Emmanouil S. [Cardiology Division, Dallas VA Medical Center and UT Southwestern Medical Center, Dallas, TX (United States); Lombardi, William L. [University of Washington, Cardiology Division, Seattle, WA (United States)

    2017-06-15

    The purpose of this study was to demonstrate the feasibility of real-time fusion of coronary computed tomography angiography (CTA) centreline and arterial wall calcification with X-ray fluoroscopy during chronic total occlusion (CTO) percutaneous coronary intervention (PCI). Patients undergoing CTO PCI were prospectively enrolled. Pre-procedural CT scans were integrated with conventional coronary fluoroscopy using prototype software. We enrolled 24 patients who underwent CTO PCI using the prototype CT fusion software, and 24 consecutive CTO PCI patients without CT guidance served as a control group. Mean age was 66 ± 11 years, and 43/48 patients were men. Real-time CTA fusion during CTO PCI provided additional information regarding coronary arterial calcification and tortuosity that generated new insights into antegrade wiring, antegrade dissection/reentry, and retrograde wiring during CTO PCI. Overall CTO success rates and procedural outcomes remained similar between the two groups, despite a trend toward higher complexity in the fusion CTA group. This study demonstrates that real-time automated co-registration of coronary CTA centreline and calcification onto live fluoroscopic images is feasible and provides new insights into CTO PCI, and in particular, antegrade dissection reentry-based CTO PCI. (orig.)

  10. FDG-PET/CT detection of very early breast cancer in women with breast microcalcification lesions found in mammography screening

    International Nuclear Information System (INIS)

    Peng, Nang-Jing; Chou, Chen-Pin; Pan, Huay-Ben; Chang, Tsung-Hsien; Hu, Chin; Chiu, Yu-Li; Fu, Ting-Ying; Chang, Hong-Tai

    2015-01-01

    To assess the efficacy of positron emission tomography/computed tomography with the glucose analogue 2-[ 18 F]fluoro-2-deoxy-D-glucose (FDG-PET/CT) in Taiwanese women with early breast cancer detected by mammography screening. Dual-time-point imaging of whole-body supine and breast prone scans using FDG-PET/CT were performed sequentially in the pre-operative stage. A total of 11,849 patients underwent screening mammography, of whom 1,209 (10.2%) displayed positive results. After further investigation, 54 patients underwent FDG-PET/CT. Post-operative pathology examinations revealed malignancies in 26 lesions, including invasive breast cancer in 11 cases and non-invasive breast cancer in 15 cases, as well as benign disease in 30 lesions. The FDG-PET/CT findings from the whole-body scans were positive for 9 of 11 invasive breast cancers (81.8%) and 3 of 15 non-invasive cancers (20%), and they were negative for all benign lesions. The sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of FDG-PET/CT with whole-body supine imaging were 46.2%, 100%, 100% and 68.2%, respectively. Breast prone imaging revealed another patient with ductal carcinoma in situ, increasing the sensitivity to 50%. Importantly, positive PET findings were significantly correlated with tumour histology (P = 0.006), tumour size (P = 0.039) and Ki-67 expression (P = 0.011). FDG-PET/CT with whole-body scanning demonstrated high sensitivity to invasive breast cancer, limited sensitivity to non-invasive breast cancer, and high specificity for breast cancer. FDG-PET/CT might be useful for differentiating tumour invasiveness. However, the good PPV but poor NPV do not allow the physician to discard the biopsy.

  11. Structural characterization of titanium porous foams by gamma rays transmission and X ray microtomography

    International Nuclear Information System (INIS)

    Moreira, Anderson C.; Appoloni, Carlos R.

    2007-01-01

    The advance in porous media studies and the consequent progresses of their applications in medicine, petroleum industry, metallurgy and others, rises the prominence of the area among scientists. This fact increases the research on different characterization techniques of porous materials. In this work three Titanium foams were analyzed by gamma rays transmission (GRT), and one of these was also analyzed by X-ray microtomography (μ-CT). The GRT experimental set consisted by a 2'' x 2''. NaI(Tl) detector, a 241-Am radioactive source (59,53 keV, 100 mCi) and a standard gamma spectrometry electronic chain. In the μ-CT technique it was used a SKYSCAN 1172 scanner consisting of an X-ray tube (20-100 kV and 0-250 μA), a CCD detector and a proper mechanic system for sample and detector movement. The system may reach ∼0.8 μm image resolution. Images of 815 slices of the sample were generated and analyzed by the IMAGO software. It permitted the determination of geometrical parameters like pore size distribution, total porosity and autocorrelation function. The analysis of data, obtained by both techniques, showed that porous media are homogeneous in the reached resolutions (1 mm to GRT and 5 μm to μ-CT). The average total porosities determined by GRT for each sample were φ 1 =53,47±0,36%, φ 2 =55,95±0,23% and φ 3 =56,80±0,56%, and the determined by μ-CT was φ 1 =53,47±0,36%. The porosity data of the Ti-1 sample shows good agreement from both techniques. The pore size distribution, from μ-CT technique of the Ti-1 sample, showed that 57% of porous phase have porous with radius in 20 to 80 μm range. (author)

  12. High speed imaging of dynamic processes with a switched source x-ray CT system

    International Nuclear Information System (INIS)

    Thompson, William M; Lionheart, William R B; Morton, Edward J; Cunningham, Mike; Luggar, Russell D

    2015-01-01

    Conventional x-ray computed tomography (CT) scanners are limited in their scanning speed by the mechanical constraints of their rotating gantries and as such do not provide the necessary temporal resolution for imaging of fast-moving dynamic processes, such as moving fluid flows. The Real Time Tomography (RTT) system is a family of fast cone beam CT scanners which instead use multiple fixed discrete sources and complete rings of detectors in an offset geometry. We demonstrate the potential of this system for use in the imaging of such high speed dynamic processes and give results using simulated and real experimental data. The unusual scanning geometry results in some challenges in image reconstruction, which are overcome using algebraic iterative reconstruction techniques and explicit regularisation. Through the use of a simple temporal regularisation term and by optimising the source firing pattern, we show that temporal resolution of the system may be increased at the expense of spatial resolution, which may be advantageous in some situations. Results are given showing temporal resolution of approximately 500 µs with simulated data and 3 ms with real experimental data. (paper)

  13. SU-F-I-01: Normalized Mean Glandular Dose Values for Dedicated Breast CT Using Realistic Breast-Shaped Phantoms

    Energy Technology Data Exchange (ETDEWEB)

    Hernandez, A [Department of Radiology, Biomedical Engineering Graduate Group, University of California Davis, Sacramento, CA (United States); Boone, J [Departments of Radiology and Biomedical Engineering, Biomedical Engeering Graduate Group, University of California Davis, Sacramento, CA (United States)

    2016-06-15

    Purpose: To estimate normalized mean glandular dose values for dedicated breast CT (DgN-CT) using breast CT-derived phantoms and compare to estimations using cylindrical phantoms. Methods: Segmented breast CT (bCT) volume data sets (N=219) were used to measure effective diameter profiles and were grouped into quintiles by volume. The profiles were averaged within each quintile to represent the range of breast sizes found clinically. These profiles were then used to generate five voxelized computational phantoms (V1, V2, V3, V4, V5 for the small to large phantom sizes, respectively), and loaded into the MCNP6 lattice geometry to simulate normalized mean glandular dose coefficients (DgN-CT) using the system specifications of the Doheny-prototype bCT scanner in our laboratory. The DgN-CT coefficients derived from the bCT-derived breast-shaped phantoms were compared to those generated using a simpler cylindrical phantom using a constant volume, and the following constraints: (1) Length=1.5*radius; (2) radius determined at chest wall (Rcw), and (3) radius determined at the phantom center-of-mass (Rcm). Results: The change in Dg-NCT coefficients averaged across all phantom sizes, was - 0.5%, 19.8%, and 1.3%, for constraints 1–3, respectively. This suggests that the cylindrical assumption is a good approximation if the radius is taken at the breast center-of-mass, but using the radius at the chest wall results in an underestimation of the glandular dose. Conclusion: The DgN-CT coefficients for bCT-derived phantoms were compared against the assumption of a cylindrical phantom and proved to be essentially equivalent when the cylinder radius was set to r=1.5/L or Rcm. While this suggests that for dosimetry applications a patient’s breast can be approximated as a cylinder (if the correct radius is applied), this assumes a homogenous composition of breast tissue and the results may be different if the realistic heterogeneous distribution of glandular tissue is considered

  14. The journey: from X-rays to PET-MRI

    International Nuclear Information System (INIS)

    Sheikh, Tariq Hussain

    2010-01-01

    Full text: Medical imaging has undergone remarkable evolution over the past century. Since the discovery of the X-rays by (Wilhelm Conrad Roentgen), static emission tomography (Hal Anger) computed tomography (Godfrey Hounsfield and Alan Cormack), and magnetic resonance imaging (Paul Lauterbur and Peter Mansfield) there have been many other important discoveries and technical developments that have culminated in our current sophisticated multi-modality imaging systems. Nobel Prizes have been given for the discoveries of radioactivity (Marie Curie, Pierre Curie, and Henri Becquerel in 1903) and the positron (Carl Anderson in 1936) and for technical developments such as the radiotracer concept (George De Hevesy in 1943). Positron emission detection systems have developed since their first use in the 1950s to the high-resolution, high-sensitivity tomographic devices that we have today. In keeping pace with these milestones in the evolution of medical imaging, positron emission tomography (PET), and more recently integrated positron emission tomography-computed tomography (PET-CT), have now emerged not only as important research tools but also as significant diagnostic imaging systems in clinical medicine. The use of multi-modality imaging systems and 'smart' specific imaging agents will achieve the key task of accurate diagnosis, treatment evaluation, surveillance, and prognosis in individual patients. PET-CT instrumentation has continued to evolve rapidly, especially over the last decade A PET scanner is combined with a CT scanner into a single machine. The PET and CT components are mounted on the same aluminium support with the CT on the front and PET at the back. Metabolic information is obtained from the PET scanner (emission of annihilation photons) and anatomic information is obtained from the CT scan (transmission of X-Rays). In addition, the CT scan can be used to provide information needed for attenuation correction. The current generation of PET-CT scanners

  15. X-ray fluorescence microtomography analyzing prostate tissues

    International Nuclear Information System (INIS)

    Pereira, Gabriela R.; Rocha, Henrique S.; Calza, Cristiane; Lopes, Ricardo T.

    2009-01-01

    The objective of this work is to determine the elemental distribution map in reference samples and prostate tissue samples using X-Ray Fluorescence Microtomography (XRFCT) in order to verify concentrations of certain elements correlated with characteristics observed by the transmission microtomography. The experiments were performed at the X-Ray Fluorescence Facility of the Brazilian Synchrotron Light Laboratory. A quasi-monochromatic beam produced by a multilayer monochromator was used as an incident beam. The transmission CT images were reconstructed using filtered-back-projection algorithm, and the XRFCT images were reconstructed using filtered-back-projection algorithm with absorption corrections. (author)

  16. The influence of mammographic X-ray spectra on absorbed energy distribution in breast: Monte Carlo simulation studies

    International Nuclear Information System (INIS)

    Delis, H.; Spyrou, G.; Tzanakos, G.; Panayiotakis, G.

    2005-01-01

    A mathematical model, based on Monte Carlo simulation, is proposed for deriving absorbed energy and dose distribution in mammography utilizing a mathematical water-like phantom. The model was validated for its accuracy against experimental and published data. The main factor discriminating absorbed energy distribution characteristics among different mammographic techniques was considered the X-ray spectrum. The absorbed energy distribution inside the phantom was investigated via percentage depth dose and isodose curves. The influence of the factors affecting X-ray spectrum (tube voltage, anode material, filter material and thickness) on absorbed energy distribution was examined. The hardness of the beam, due to increase of tube voltage or filtration, was found to be the major factor affecting absorbed energy distribution inside the phantom. In general, Mo and W anode systems demonstrated superior dosimetric characteristics against those of W-Mo or Rh. The model presented can be used for estimating absolute and relative breast dose values and their spatial distributions

  17. Semi-automated scoring of pulmonary emphysema from X-ray CT: Trainee reproducibility and accuracy

    Energy Technology Data Exchange (ETDEWEB)

    Owrangi, Amir M., E-mail: aowrangi@robarts.ca [Imaging Research Laboratories, Robarts Research Institute, London (Canada); Entwistle, Brandon, E-mail: Brandon.Entwistle@londonhospitals.ca; Lu, Andrew, E-mail: Andrew.Lu@londonhospitals.ca; Chiu, Jack, E-mail: Jack.Chiu@londonhospitals.ca; Hussain, Nabil, E-mail: Nabil.Hussain@londonhospitals.ca; Etemad-Rezai, Roya, E-mail: Roya.EtemadRezai@lhsc.on.ca; Parraga, Grace, E-mail: gparraga@robarts.ca [Imaging Research Laboratories, Robarts Research Institute, London (Canada); Graduate Program in Biomedical Engineering, Department of Medical Imaging, Department of Medical Biophysics, The University of Western Ontario, London (Canada)

    2013-11-01

    Objective: We developed a semi-automated tool to quantify emphysema from thoracic X-ray multi-detector (64-slice) computed tomography (CT) for training purposes and multi-reader studies. Materials and Methods: Thoracic X-ray CT was acquired in 93 ex-smokers, who were evaluated by six trainees with little or no expertise (trainees) and a single experienced thoracic radiologist (expert). A graphic user interface (GUI) was developed for emphysema quantification based on the percentile of lung where a score of 0 = no abnormalities, 1 = 1–25%, 2 = 26–50%, 3 = 51–75% and 4 = 76–100% for each lung side/slice. Trainees blinded to subject characteristics scored randomized images twice; accuracy was determined by comparison to expert scores, density histogram 15th percentile (HU{sub 15}), relative area at −950 HU (RA{sub 950}), low attenuation clusters at −950 HU (LAC{sub 950}), −856 HU (LAC{sub 856}) and the diffusing capacity for carbon monoxide (DL{sub CO%pred}). Intra- and inter-observer reproducibility was evaluated using coefficients-of-variation (COV), intra-class (ICC) and Pearson correlations. Results: Trainee–expert correlations were significant (r = 0.85–0.97, p < 0.0001) and a significant trainee bias (0.15 ± 0.22) was observed. Emphysema score was correlated with RA{sub 950} (r = 0.88, p < 0.0001), HU{sub 15} (r = −0.77, p < 0.0001), LAC{sub 950} (r = 0.76, p < 0.0001), LAC{sub 856} (r = 0.74, p = 0.0001) and DL{sub CO%pred} (r = −0.71, p < 0.0001). Intra-observer reproducibility (COV = 4–27%; ICC = 0.75–0.94) was moderate to high for trainees; intra- and inter-observer COV were negatively and non-linearly correlated with emphysema score. Conclusion: We developed a GUI for rapid and interactive emphysema scoring that allows for comparison of multiple readers with clinical and radiological standards.

  18. Metal artifact removal (MAR) analysis for the security inspections using the X-ray computed tomography

    International Nuclear Information System (INIS)

    Cho, Hyo Sung; Woo, Tae Ho; Park, Chul Kyu

    2016-01-01

    Using the metal artifact property, it is analyzed for the X-ray computed tomography (CT) in the aspect of the security on the examined places like airport and surveillance areas. Since the importance of terror prevention strategy has been increased, the security application of X-ray CT has the significant remark. One shot X-ray image has the limitation to find out the exact shape to property in the closed box, which could be solved by the CT scanning without the tearing off the box in this work. Cleaner images can be obtained by the advanced technology if the CT scanning is utilized in the security purposes on the secured areas. A metal sample is treated by the metal artifact removal (MAR) method for the enhanced image. The mimicked explosive is experimented for the imaging processing application where the cleaner one is obtained. The procedure is explained and the further study is discussed. - Highlights: • The X-ray image is used for the nuclear inspections which are made by the image processing. • Nuclear security has been performed for counterterrorism. • Easy and fast inspections are needed. • Metal shows the characteristics of the noises.

  19. Impact of intense x-ray pulses on a NaI(Tl)-based gamma camera

    NARCIS (Netherlands)

    Koppert, Wilco J C; van der Velden, Sandra; Steenbergen, J H Leo; de Jong, Hugo W A M

    2018-01-01

    INTRODUCTION: In SPECT/CT systems X-ray and -ray imaging is performed sequentially. Simultaneous acquisition may have advantages, for instance in interventional settings. However, this may expose a gamma camera to relatively high X-ray doses and deteriorate its functioning. We studied the NaI(Tl)

  20. Cancer risk analysis among medical diagnostic X-ray workers in China

    International Nuclear Information System (INIS)

    Wang Jixian; Li Benxiao; Gao Zhiwei; Xu Jun; Zhang Jingyuan; Aoyama, Takashi; Sugahara, Tsutomu

    1997-01-01

    To provide the evidence and related rules of human malignant tumors produced by prolonged exposure to low level ionizing radiation. The cancer incidence (1950-1990) among 27011 medical diagnostic X-ray workers was compared with that among 25782 other medical specialists employed between 1950 and 1980 in China by means of O/E system. A significantly elevated risk of cancer was seen among diagnostic x-ray workers (RR = 1.1,95%Cl:1.0-1.2,P <0.05). Significantly elevated risks were seen for leukemia and cancers of skin, female breast, liver and esophagus, the RR being 2.3, 5.0, 1.6, 1.3 and 4.4 respectively. The RR for leukemia was higher for X-ray workers who began employment before 1970 and also for those who were young when employment began. The patterns of risk associated with length of service and with age and calendar year of initial employment suggest that the excesses of leukemia, skin cancer and female breast cancer were due to occupational exposure to X-rays. The ERR and EAR for leukemia and solid cancer were calculated roughly. (author)

  1. Rod displacement measurements by x-ray CT and its impact on thermal-hydraulics in tight-lattice rod bundle (Joint research)

    International Nuclear Information System (INIS)

    Mitsutake, Toru; Misawa, Takeharu; Kureta, Masatoshi; Akimoto, Hajime

    2005-06-01

    In tight-lattice simulated rod bundles with about 1 mm gap between rods, a rod displacement might affect thermal-hydraulic characteristics since the displacement has a strong impact on the flow area change along the heated section. It should be important to estimate how large the rod position displacement could quantitatively affect critical power for the tight-lattice rod bundle from the point of improvement of prediction capability of subchannel analysis. In the present study, the inside-structure observation of the simulated seven-rod bundle of Reduced Moderation Water Reactor (RMWR) was made through the whole length of the test assembly. Based on the measured rod position data, the relation between the rod position displacement and the heat transfer characteristics was investigated experimentally and through the two kinds of subchannel analysis, the nominal rod position case and the measured rod position case, the effect on the predicted critical power was estimated. The high-energy X-ray computer tomograph (CT) of Fuels Monitoring Facilities (FMF) at the O-arai Engineering Center in Japan Nuclear Cycle Institute (JNC) was applied for the inside-structure observation of the test assembly. The CT view of the cross sections within the test assembly assured the hexagonal rod position arrangement was almost the same as expected by design. The measured data with the X-ray CT facility showed that all rod displacements were small, 0.5 millimeters at maximum and 0.2 millimeters in average. In the heat transfer experiments for the seven-rod bundle, the boiling transition (BT) position and the rod surface temperature behavior was measured. All thermocouples on the center rod downstream from the BT-onset axial height showed almost simultaneous temperature increase due to BT. And the thermocouples located on the same axial heights showed quite similar time-variation behaviors in the vapor cooling heat transfer regime. These results demonstrated the effect of the

  2. MO-F-CAMPUS-I-02: Accuracy in Converting the Average Breast Dose Into the Mean Glandular Dose (MGD) Using the F-Factor in Cone Beam Breast CT- a Monte Carlo Study Using Homogeneous and Quasi-Homogeneous Phantoms

    Energy Technology Data Exchange (ETDEWEB)

    Lai, C; Zhong, Y; Wang, T; Shaw, C [UT MD Anderson Cancer Center, Houston, TX (United States)

    2015-06-15

    Purpose: To investigate the accuracy in estimating the mean glandular dose (MGD) for homogeneous breast phantoms by converting from the average breast dose using the F-factor in cone beam breast CT. Methods: EGSnrc-based Monte Carlo codes were used to estimate the MGDs. 13-cm in diameter, 10-cm high hemi-ellipsoids were used to simulate pendant-geometry breasts. Two different types of hemi-ellipsoidal models were employed: voxels in quasi-homogeneous phantoms were designed as either adipose or glandular tissue while voxels in homogeneous phantoms were designed as the mixture of adipose and glandular tissues. Breast compositions of 25% and 50% volume glandular fractions (VGFs), defined as the ratio of glandular tissue voxels to entire breast voxels in the quasi-homogeneous phantoms, were studied. These VGFs were converted into glandular fractions by weight and used to construct the corresponding homogeneous phantoms. 80 kVp x-rays with a mean energy of 47 keV was used in the simulation. A total of 109 photons were used to image the phantoms and the energies deposited in the phantom voxels were tallied. Breast doses in homogeneous phantoms were averaged over all voxels and then used to calculate the MGDs using the F-factors evaluated at the mean energy of the x-rays. The MGDs for quasi-homogeneous phantoms were computed directly by averaging the doses over all glandular tissue voxels. The MGDs estimated for the two types of phantoms were normalized to the free-in-air dose at the iso-center and compared. Results: The normalized MGDs were 0.756 and 0.732 mGy/mGy for the 25% and 50% VGF homogeneous breasts and 0.761 and 0.733 mGy/mGy for the corresponding quasi-homogeneous breasts, respectively. The MGDs estimated for the two types of phantoms were similar within 1% in this study. Conclusion: MGDs for homogeneous breast models may be adequately estimated by converting from the average breast dose using the F-factor.

  3. MO-F-CAMPUS-I-02: Accuracy in Converting the Average Breast Dose Into the Mean Glandular Dose (MGD) Using the F-Factor in Cone Beam Breast CT- a Monte Carlo Study Using Homogeneous and Quasi-Homogeneous Phantoms

    International Nuclear Information System (INIS)

    Lai, C; Zhong, Y; Wang, T; Shaw, C

    2015-01-01

    Purpose: To investigate the accuracy in estimating the mean glandular dose (MGD) for homogeneous breast phantoms by converting from the average breast dose using the F-factor in cone beam breast CT. Methods: EGSnrc-based Monte Carlo codes were used to estimate the MGDs. 13-cm in diameter, 10-cm high hemi-ellipsoids were used to simulate pendant-geometry breasts. Two different types of hemi-ellipsoidal models were employed: voxels in quasi-homogeneous phantoms were designed as either adipose or glandular tissue while voxels in homogeneous phantoms were designed as the mixture of adipose and glandular tissues. Breast compositions of 25% and 50% volume glandular fractions (VGFs), defined as the ratio of glandular tissue voxels to entire breast voxels in the quasi-homogeneous phantoms, were studied. These VGFs were converted into glandular fractions by weight and used to construct the corresponding homogeneous phantoms. 80 kVp x-rays with a mean energy of 47 keV was used in the simulation. A total of 109 photons were used to image the phantoms and the energies deposited in the phantom voxels were tallied. Breast doses in homogeneous phantoms were averaged over all voxels and then used to calculate the MGDs using the F-factors evaluated at the mean energy of the x-rays. The MGDs for quasi-homogeneous phantoms were computed directly by averaging the doses over all glandular tissue voxels. The MGDs estimated for the two types of phantoms were normalized to the free-in-air dose at the iso-center and compared. Results: The normalized MGDs were 0.756 and 0.732 mGy/mGy for the 25% and 50% VGF homogeneous breasts and 0.761 and 0.733 mGy/mGy for the corresponding quasi-homogeneous breasts, respectively. The MGDs estimated for the two types of phantoms were similar within 1% in this study. Conclusion: MGDs for homogeneous breast models may be adequately estimated by converting from the average breast dose using the F-factor

  4. Children's (Pediatric) CT (Computed Tomography)

    Medline Plus

    Full Text Available ... many ways, CT scanning is like other x-ray examinations. X-rays are a form of radiation, like light ... denotes child-specific content. Related Articles and Media Radiation Dose in X-Ray and CT Exams Contrast Materials Anesthesia Safety ...

  5. Three-dimensional modeling and simulation of asphalt concrete mixtures based on X-ray CT microstructure images

    Directory of Open Access Journals (Sweden)

    Hainian Wang

    2014-02-01

    Full Text Available X-ray CT (computed tomography was used to scan asphalt mixture specimen to obtain high resolution continuous cross-section images and the meso-structure. According to the theory of three-dimensional (3D reconstruction, the 3D reconstruction algorithm was investigated in this paper. The key to the reconstruction technique is the acquisition of the voxel positions and the relationship between the pixel element and node. Three-dimensional numerical model of asphalt mixture specimen was created by a self-developed program. A splitting test was conducted to predict the stress distributions of the asphalt mixture and verify the rationality of the 3D model.

  6. Evaluation of X-ray shielding performance of protective aprons

    International Nuclear Information System (INIS)

    Kumagai, Michitomo; Shintani, Mitsuo; Kuranishi, Makoto

    1999-01-01

    Lead equivalent, which offers protection against x-rays, is rated with a 100 kV tube voltage in Japanese Industrial Standard (JIS) Z 4501-1988, Testing method of lead equivalent for x-ray protective devices.'' However, the actual tube voltage in general diagnostic examinations (normal to special radiography; including computed tomography, CT) is 50 to 150 kV. Therefore, we measured whether the performance of current lead aprons (three products) and protective aprons using composite materials (two products) changes at 60 to 141 kV of tube voltage. Furthermore, we evaluated x-ray shielding performance by measuring the transmission ratio of scattered x-rays. The lead equivalent of two currently used lead aprons was almost the same at all voltages. However, in one currently used lead apron and both protective aprons made of composite materials, lead equivalent decreased rapidly when tube voltage exceeded 100 kV. The transmission ratio of scattered x-rays increased with increasing tube voltage in all of the protective aprons examined. Further, in all aprons examined, the transmission ratio of scattered x-rays declined with widening of the scatter angle. As mentioned above, the x-ray shielding performance of some x-ray protective aprons suddenly decreased at tube voltages over 100 kV. Thus the performance of x-ray protective aprons should be published, and JIS Z 4501 needs to be revised. (author)

  7. Evaluation of X-ray shielding performance of protective aprons

    Energy Technology Data Exchange (ETDEWEB)

    Kumagai, Michitomo; Shintani, Mitsuo; Kuranishi, Makoto [Toyama Medical and Pharmaceutical Univ. (Japan). Hospital

    1999-04-01

    Lead equivalent, which offers protection against x-rays, is rated with a 100 kV tube voltage in Japanese Industrial Standard (JIS) Z 4501-1988, Testing method of lead equivalent for x-ray protective devices.`` However, the actual tube voltage in general diagnostic examinations (normal to special radiography; including computed tomography, CT) is 50 to 150 kV. Therefore, we measured whether the performance of current lead aprons (three products) and protective aprons using composite materials (two products) changes at 60 to 141 kV of tube voltage. Furthermore, we evaluated x-ray shielding performance by measuring the transmission ratio of scattered x-rays. The lead equivalent of two currently used lead aprons was almost the same at all voltages. However, in one currently used lead apron and both protective aprons made of composite materials, lead equivalent decreased rapidly when tube voltage exceeded 100 kV. The transmission ratio of scattered x-rays increased with increasing tube voltage in all of the protective aprons examined. Further, in all aprons examined, the transmission ratio of scattered x-rays declined with widening of the scatter angle. As mentioned above, the x-ray shielding performance of some x-ray protective aprons suddenly decreased at tube voltages over 100 kV. Thus the performance of x-ray protective aprons should be published, and JIS Z 4501 needs to be revised. (author)

  8. Possible Radiation-Induced Damage to the Molecular Structure of Wooden Artifacts Due to Micro-Computed Tomography, Handheld X-Ray Fluorescence, and X-Ray Photoelectron Spectroscopic Techniques

    Directory of Open Access Journals (Sweden)

    Madalena Kozachuk

    2016-05-01

    Full Text Available This study was undertaken to ascertain whether radiation produced by X-ray photoelectron spectroscopy (XPS, micro-computed tomography (μCT and/or portable handheld X-ray fluorescence (XRF equipment might damage wood artifacts during analysis. Changes at the molecular level were monitored by Fourier transform infrared (FTIR analysis. No significant changes in FTIR spectra were observed as a result of μCT or handheld XRF analysis. No substantial changes in the collected FTIR spectra were observed when XPS analytical times on the order of minutes were used. However, XPS analysis collected over tens of hours did produce significant changes in the FTIR spectra.

  9. Breast dose in mammography is about 30% lower when realistic heterogeneous glandular distributions are considered

    Energy Technology Data Exchange (ETDEWEB)

    Hernandez, Andrew M., E-mail: amhern@ucdavis.edu [Biomedical Engineering Graduate Group, University of California Davis, Sacramento, California 95817 (United States); Seibert, J. Anthony; Boone, John M. [Departments of Radiology and Biomedical Engineering, Biomedical Engineering Graduate Group, University of California Davis, Sacramento, California 95817 (United States)

    2015-11-15

    Purpose: Current dosimetry methods in mammography assume that the breast is comprised of a homogeneous mixture of glandular and adipose tissues. Three-dimensional (3D) dedicated breast CT (bCT) data sets were used previously to assess the complex anatomical structure within the breast, characterizing the statistical distribution of glandular tissue in the breast. The purpose of this work was to investigate the effect of bCT-derived heterogeneous glandular distributions on dosimetry in mammography. Methods: bCT-derived breast diameters, volumes, and 3D fibroglandular distributions were used to design realistic compressed breast models comprised of heterogeneous distributions of glandular tissue. The bCT-derived glandular distributions were fit to biGaussian functions and used as probability density maps to assign the density distributions within compressed breast models. The MCNPX 2.6.0 Monte Carlo code was used to estimate monoenergetic normalized mean glandular dose “DgN(E)” values in mammography geometry. The DgN(E) values were then weighted by typical mammography x-ray spectra to determine polyenergetic DgN (pDgN) coefficients for heterogeneous (pDgN{sub hetero}) and homogeneous (pDgN{sub homo}) cases. The dependence of estimated pDgN values on phantom size, volumetric glandular fraction (VGF), x-ray technique factors, and location of the heterogeneous glandular distributions was investigated. Results: The pDgN{sub hetero} coefficients were on average 35.3% (SD, 4.1) and 24.2% (SD, 3.0) lower than the pDgN{sub homo} coefficients for the Mo–Mo and W–Rh x-ray spectra, respectively, across all phantom sizes and VGFs when the glandular distributions were centered within the breast phantom in the coronal plane. At constant breast size, increasing VGF from 7.3% to 19.1% lead to a reduction in pDgN{sub hetero} relative to pDgN{sub homo} of 23.6%–27.4% for a W–Rh spectrum. Displacement of the glandular distribution, at a distance equal to 10% of the

  10. Gamma-ray CT from incomplete projections for two-phase pipe flow.

    Science.gov (United States)

    Xin, S; Wang, H X

    2017-02-01

    A low-energy low-dose γ-ray computed tomography (CT) system used in the gas-liquid two-phase pipe flow measurement has been studied at Tianjin University in recent years. The γ-ray CT system, having a third-generation X-ray CT scanning configuration, is comprised of one 300mCi 241 Am source and 17 CdZnTe detector units and achieves a spatial image resolution of about 7 mm. It is primarily intended to measure the two-phase pipe flow and provide improvement suggestions for industrial CT system. Recently we improve the design for image reconstruction from incomplete projection to optimize the scanning parameters and reduce the radiation dose. First, tomographic problem from incomplete projections is briefly described. Next, a system structure and a hardware circuit design are listed and explained, especially on time parameter setting of the pulse shaper. And then a detailed system analysis is provided in Section II, mainly focusing on spatial resolution, temporal resolution, system noise, and imaging algorithm. Finally, we carry on necessary static and dynamic experiments in a full scan (360°) and two sets of partial scan reconstruction tests to determine the feasibility of this γ-ray CT system for reconstructing the images from insufficient projections. And based on an A-variable algebraic reconstruction technique method, a specially designed algorithm, we evaluate the system performance and noise level of this CT system working quantitatively and qualitatively. Results of dynamic test indicate that the acceptable results can be acquired using a multi-source γ-ray CT system with the same parameters when the flow rate is less than 0.04 m/s and the imaging speed is slower than 33 frames/s.

  11. Scan Quality and Entrance Skin Dose in Thoracic CT: A Comparison between Bismuth Breast Shield and Posteriorly Centered Partial CT Scans

    International Nuclear Information System (INIS)

    Tappouni, Rafel; Mathers, Bradley

    2012-01-01

    Objectives. To compare the effectiveness of the bismuth breast shield and partial CT scan in reducing entrance skin dose and to evaluate the effect of the breast shield on image quality (IQ). Methods. Nanodots were placed on an adult anthropomorphic phantom. Standard chest CT, CT with shield, and partial CT were performed. Nanodot readings and effective doses were recorded. 50 patients with chest CTs obtained both with and without breast shields were reviewed. IQ was evaluated by two radiologists and by measuring Hounsfield units (HUs) and standard deviation (SD) of HU in anterior subcutaneous region. Results. Breast shield and the partial CT scans reduced radiation to the anterior chest by 38% and 16%, respectively. Partial CT increased dose to the posterior chest by 37% and effective dose by 8%. Change in IQ in shield CT was observed in the anterior chest wall. Significant change in IQ was observed in 5/50 cases. The shield caused an increase of 20 HU (P = 0.021) and a 1.86 reduction in SD of HU (P = 0.027) in the anterior compared to posterior subcutaneous regions. Summary. Bismuth breast shield is more effective than the partial CT in reducing entrance skin dose while maintaining image quality

  12. X-ray filter for x-ray powder diffraction

    Science.gov (United States)

    Sinsheimer, John Jay; Conley, Raymond P.; Bouet, Nathalie C. D.; Dooryhee, Eric; Ghose, Sanjit

    2018-01-23

    Technologies are described for apparatus, methods and systems effective for filtering. The filters may comprise a first plate. The first plate may include an x-ray absorbing material and walls defining first slits. The first slits may include arc shaped openings through the first plate. The walls of the first plate may be configured to absorb at least some of first x-rays when the first x-rays are incident on the x-ray absorbing material, and to output second x-rays. The filters may comprise a second plate spaced from the first plate. The second plate may include the x-ray absorbing material and walls defining second slits. The second slits may include arc shaped openings through the second plate. The walls of the second plate may be configured to absorb at least some of second x-rays and to output third x-rays.

  13. Development of a CT simulator and its current clinical status

    International Nuclear Information System (INIS)

    Nagata, Yasushi; Hiraoka, Masahiro; Nishidai, Takehiro

    2000-01-01

    A computed tomography (CT) simulator is defined as a radiotherapy treatment planning (RTP) system which consists of an X-ray CT scanner, a treatment planning computer, and a marking projector. The developmental process from the introduction of an X-ray CT to a CT simulator is described in this article. After the development of the first CT simulator, several new developments in CT scanners, marking projector, and data transfer networks were introduced. CT simulator systems have recently become more widely available and are produced by more than 8 commercial companies. The advantages of a CT simulator are a shortening of overall RTP time, increased accuracy of RTP, and the possibility of 3-D conformal therapy. Clinical experience with CT simulators has accumulated over the past 10 years. Site-specific trials have been undertaken, and the clinical usefulness of CT simulators for breast cancer, maxillary cancer, nasopharyngeal cancer, orbital tumor, lung cancer and other tumors has been demonstrated. The use of CT simulators has resulted in a decrease in the complications of radiotherapy. It also appears to be an essential tool for dose escalation studies, which may permit increased local control of certain tumors. In Japan, CT simulators were in use at more than 134 (20%) of 682 radiotherapy institutions as of the end of 1998. In the USA, CT simulators were present in more than 12% of 1,542 radiotherapy centers in 1994. The CT simulator is now an essential RTP system for advanced radiotherapy. (author)

  14. Ultralow dose CT for pulmonary nodule detection with chest X-ray equivalent dose - a prospective intra-individual comparative study

    Energy Technology Data Exchange (ETDEWEB)

    Messerli, Michael [University Zurich, Department of Nuclear Medicine, University Hospital Zurich, Zurich (Switzerland); Cantonal Hospital St. Gallen, Division of Radiology and Nuclear Medicine, St. Gallen (Switzerland); Kluckert, Thomas; Knitel, Meinhard; Desbiolles, Lotus; Bauer, Ralf W.; Wildermuth, Simon [Cantonal Hospital St. Gallen, Division of Radiology and Nuclear Medicine, St. Gallen (Switzerland); Waelti, Stephan [Cantonal Hospital St. Gallen, Division of Radiology and Nuclear Medicine, St. Gallen (Switzerland); University of Montreal, Department of Radiology, CHU Sainte-Justine, Montreal, Quebec (Canada); Rengier, Fabian [University Hospital Heidelberg, Department of Diagnostic and Interventional Radiology, Heidelberg (Germany); Warschkow, Rene [Cantonal Hospital St. Gallen, Department of Surgery, St. Gallen (Switzerland); Alkadhi, Hatem [University Zurich, Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Zurich (Switzerland); Leschka, Sebastian [Cantonal Hospital St. Gallen, Division of Radiology and Nuclear Medicine, St. Gallen (Switzerland); University Zurich, Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, Zurich (Switzerland)

    2017-08-15

    To prospectively evaluate the accuracy of ultralow radiation dose CT of the chest with tin filtration at 100 kV for pulmonary nodule detection. 202 consecutive patients undergoing clinically indicated chest CT (standard dose, 1.8 ± 0.7 mSv) were prospectively included and additionally scanned with an ultralow dose protocol (0.13 ± 0.01 mSv). Standard dose CT was read in consensus by two board-certified radiologists to determine the presence of lung nodules and served as standard of reference (SOR). Two radiologists assessed the presence of lung nodules and their locations on ultralow dose CT. Sensitivity and specificity of the ultralow dose protocol was compared against the SOR, including subgroup analyses of different nodule sizes and types. A mixed effects logistic regression was used to test for independent predictors for sensitivity of pulmonary nodule detection. 425 nodules (mean diameter 3.7 ± 2.9 mm) were found on SOR. Overall sensitivity for nodule detection by ultralow dose CT was 91%. In multivariate analysis, nodule type, size and patients BMI were independent predictors for sensitivity (p < 0.001). Ultralow dose chest CT at 100 kV with spectral shaping enables a high sensitivity for the detection of pulmonary nodules at exposure levels comparable to plain film chest X-ray. (orig.)

  15. Multi-mounted X-ray cone-beam computed tomography

    Science.gov (United States)

    Fu, Jian; Wang, Jingzheng; Guo, Wei; Peng, Peng

    2018-04-01

    As a powerful nondestructive inspection technique, X-ray computed tomography (X-CT) has been widely applied to clinical diagnosis, industrial production and cutting-edge research. Imaging efficiency is currently one of the major obstacles for the applications of X-CT. In this paper, a multi-mounted three dimensional cone-beam X-CT (MM-CBCT) method is reported. It consists of a novel multi-mounted cone-beam scanning geometry and the corresponding three dimensional statistical iterative reconstruction algorithm. The scanning geometry is the most iconic design and significantly different from the current CBCT systems. Permitting the cone-beam scanning of multiple objects simultaneously, the proposed approach has the potential to achieve an imaging efficiency orders of magnitude greater than the conventional methods. Although multiple objects can be also bundled together and scanned simultaneously by the conventional CBCT methods, it will lead to the increased penetration thickness and signal crosstalk. In contrast, MM-CBCT avoids substantially these problems. This work comprises a numerical study of the method and its experimental verification using a dataset measured with a developed MM-CBCT prototype system. This technique will provide a possible solution for the CT inspection in a large scale.

  16. CT perfusion: the clinical application in the diagnosis of breast diseases

    International Nuclear Information System (INIS)

    Ye Zhaoxiang; Song Xiuyu; Xiao Jianyu

    2005-01-01

    Objective: To assess the CT perfusion characteristics of breast diseases. Methods: Twenty-two patients with fibroadenoma accompanied with cyst (n=1), other benign diseases (n=5), and carcinomas (n=16) underwent breast CT scan. CT perfusion was performed for breast with cine mode (0.5 s per rotation ) on a multislice CT scanner, 120 kV, 60 mA, 4 x 5 mm collimation. Contrast injection was done by using 50 ml nonionic contrast agent (300 mg I/ml ), at a flow rate of 4 ml/s with a power injector, 5 seconds delay, and data acquisition lasted for 50 seconds. These data were processed on a GE AW 4.1 workstation using perfusion 3 software package, and the mean BF, BV, MTT and PS were measured and statistically analyzed. Results: The mean BF, BV, MTT and PS in carcinoma were (33.55 ± 28.40) ml·min -1 ·100 g -1 , (5.60±3.08) ml/100 g, (15.37±6.41) s, (18.71±10.42) ml· min -1 ·100 g -1 . The mean BF, BV, MTT and PS in contralateral normal breast of carcinoma were (10.36 ± 9.44) ml·min -1 ·100 g -1 , (0.70±0.22) ml/100 g, (13.55 ± 7.27) s, (1.54±1.74) ml·min -1 ·100 g -1 . The mean BF, BV, and PS between carcinoma and normal breast were statistically significant (P 0.05). The mean BF, BV, MTT and PS in fibroadenoma were (62.57 ± 19.62) ml·min -1 ·100 g -1 , (7.37 ±1.30) ml/100 g, (11.24 ± 3.34) s, (27.63 ± 3.87) ml·min -1 ·100 g -1 . The mean PS between carcinoma and fibroadenoma was statistically significant (P 0.05). The mean BF, BV, MTT and PS in benign lesion were (7.65 ± 10.24) ml·min -1 ·100 g -1 , (1.23 ± 0.97) ml/100 g, (14.94 ± 10.68) s, (1.81 ± 1.72) ml·min -1 ·100 g -1 . The mean BF, BV, and PS between carcinoma and benign lesion were statistically significant (P 0.05). Conclusion: MSCT perfusion may reflect the features of breast diseases, and provide valuable hemodynamic message. These parameters have a value in diagnosis of breast diseases and are helpful for differential diagnosis. (authors)

  17. Large area CMOS active pixel sensor x-ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization

    Energy Technology Data Exchange (ETDEWEB)

    Zhao, Chumin; Kanicki, Jerzy, E-mail: kanicki@eecs.umich.edu [Solid-State Electronics Laboratory, Department of Electrical Engineering and Computer Science, University of Michigan, Ann Arbor, Michigan 48109 (United States); Konstantinidis, Anastasios C. [Department of Medical Physics and Biomedical Engineering, University College London, London WC1E 6BT, United Kingdom and Diagnostic Radiology and Radiation Protection, Christie Medical Physics and Engineering, The Christie NHS Foundation Trust, Manchester M20 4BX (United Kingdom); Patel, Tushita [Department of Physics, University of Virginia, Charlottesville, Virginia 22908 (United States)

    2015-11-15

    Purpose: Large area x-ray imagers based on complementary metal-oxide-semiconductor (CMOS) active pixel sensor (APS) technology have been proposed for various medical imaging applications including digital breast tomosynthesis (DBT). The low electronic noise (50–300 e{sup −}) of CMOS APS x-ray imagers provides a possible route to shrink the pixel pitch to smaller than 75 μm for microcalcification detection and possible reduction of the DBT mean glandular dose (MGD). Methods: In this study, imaging performance of a large area (29 × 23 cm{sup 2}) CMOS APS x-ray imager [Dexela 2923 MAM (PerkinElmer, London)] with a pixel pitch of 75 μm was characterized and modeled. The authors developed a cascaded system model for CMOS APS x-ray imagers using both a broadband x-ray radiation and monochromatic synchrotron radiation. The experimental data including modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE) were theoretically described using the proposed cascaded system model with satisfactory consistency to experimental results. Both high full well and low full well (LFW) modes of the Dexela 2923 MAM CMOS APS x-ray imager were characterized and modeled. The cascaded system analysis results were further used to extract the contrast-to-noise ratio (CNR) for microcalcifications with sizes of 165–400 μm at various MGDs. The impact of electronic noise on CNR was also evaluated. Results: The LFW mode shows better DQE at low air kerma (K{sub a} < 10 μGy) and should be used for DBT. At current DBT applications, air kerma (K{sub a} ∼ 10 μGy, broadband radiation of 28 kVp), DQE of more than 0.7 and ∼0.3 was achieved using the LFW mode at spatial frequency of 0.5 line pairs per millimeter (lp/mm) and Nyquist frequency ∼6.7 lp/mm, respectively. It is shown that microcalcifications of 165–400 μm in size can be resolved using a MGD range of 0.3–1 mGy, respectively. In comparison to a General Electric GEN2 prototype DBT system (at

  18. Large area CMOS active pixel sensor x-ray imager for digital breast tomosynthesis: Analysis, modeling, and characterization

    International Nuclear Information System (INIS)

    Zhao, Chumin; Kanicki, Jerzy; Konstantinidis, Anastasios C.; Patel, Tushita

    2015-01-01

    Purpose: Large area x-ray imagers based on complementary metal-oxide-semiconductor (CMOS) active pixel sensor (APS) technology have been proposed for various medical imaging applications including digital breast tomosynthesis (DBT). The low electronic noise (50–300 e − ) of CMOS APS x-ray imagers provides a possible route to shrink the pixel pitch to smaller than 75 μm for microcalcification detection and possible reduction of the DBT mean glandular dose (MGD). Methods: In this study, imaging performance of a large area (29 × 23 cm 2 ) CMOS APS x-ray imager [Dexela 2923 MAM (PerkinElmer, London)] with a pixel pitch of 75 μm was characterized and modeled. The authors developed a cascaded system model for CMOS APS x-ray imagers using both a broadband x-ray radiation and monochromatic synchrotron radiation. The experimental data including modulation transfer function, noise power spectrum, and detective quantum efficiency (DQE) were theoretically described using the proposed cascaded system model with satisfactory consistency to experimental results. Both high full well and low full well (LFW) modes of the Dexela 2923 MAM CMOS APS x-ray imager were characterized and modeled. The cascaded system analysis results were further used to extract the contrast-to-noise ratio (CNR) for microcalcifications with sizes of 165–400 μm at various MGDs. The impact of electronic noise on CNR was also evaluated. Results: The LFW mode shows better DQE at low air kerma (K a < 10 μGy) and should be used for DBT. At current DBT applications, air kerma (K a ∼ 10 μGy, broadband radiation of 28 kVp), DQE of more than 0.7 and ∼0.3 was achieved using the LFW mode at spatial frequency of 0.5 line pairs per millimeter (lp/mm) and Nyquist frequency ∼6.7 lp/mm, respectively. It is shown that microcalcifications of 165–400 μm in size can be resolved using a MGD range of 0.3–1 mGy, respectively. In comparison to a General Electric GEN2 prototype DBT system (at MGD of 2.5 m

  19. X-ray fluorescence imaging with polycapillary X-ray optics

    International Nuclear Information System (INIS)

    Yonehara, Tasuku; Yamaguchi, Makoto; Tsuji, Kouichi

    2010-01-01

    X-ray fluorescence spectrometry imaging is a powerful tool to provide information about the chemical composition and elemental distribution of a specimen. X-ray fluorescence spectrometry images were conventionally obtained by using a μ-X-ray fluorescence spectrometry spectrometer, which requires scanning a sample. Faster X-ray fluorescence spectrometry imaging would be achieved by eliminating the process of sample scanning. Thus, we developed an X-ray fluorescence spectrometry imaging instrument without sample scanning by using polycapillary X-ray optics, which had energy filter characteristics caused by the energy dependence of the total reflection phenomenon. In the present paper, we show that two independent straight polycapillary X-ray optics could be used as an energy filter of X-rays for X-ray fluorescence. Only low energy X-rays were detected when the angle between the two optical axes was increased slightly. Energy-selective X-ray fluorescence spectrometry images with projection mode were taken by using an X-ray CCD camera equipped with two polycapillary optics. It was shown that Fe Kα (6.40 keV) and Cu Kα (8.04 keV) could be discriminated for Fe and Cu foils.

  20. TU-FG-BRB-04: A New Optimization Method for Pre-Treatment Patient-Specific Stopping-Power by Combining Proton Radiography and X-Ray CT

    International Nuclear Information System (INIS)

    Collins-Fekete, C; Schulte, R; Beaulieu, L; Seco, J

    2016-01-01

    Purpose: The relative stopping power (RSP) uncertainty is the largest contributor to the range uncertainty in proton therapy. The purpose of this work is to develop a robust and systematic method that yields accurate patient specific RSPs by combining pre-treatment X-ray CT and daily proton radiography. Methods: The method is formulated as a penalized least squares optimization (PLSO) problem min(|Ax-B|). The matrix A represents the cumulative path-length crossed in each material computed by calculating proton trajectories through the X-ray CT. The material RSPs are denoted by x and B is the pRad, expressed as water equivalent thickness. The equation is solved using a convex-conic optimizer. Geant4 simulations were made to assess the feasibility of the method. RSP extracted from the Geant4 materials were used as a reference and the clinical HU-RSP curve as a comparison. The PLSO was first tested on a Gammex RMI-467 phantom. Then, anthropomorphic phantoms of the head, pelvis and lung were studied and resulting RSPs were evaluated. A pencil beam was generated in each phantom to evaluate the proton range accuracy achievable by using the optimized RSPs. Finally, experimental data of a pediatric head phantom (CIRS) were acquired using a recently completed experimental pCT scanner. Results: Numerical simulations showed precise RSP (<0.75%) for Gammex materials except low-density lung (LN-300) (1.2%). Accurate RSP have been obtained for the head (µ=−0.10%, 1.5σ=1.12%), lung (µ=−0.33, 1.5σ=1.02%) and pelvis anthropomorphic phantoms (µ=0.12, 1.5σ=0,99%). The range precision has been improved with an average R80 difference to the reference (µ±1.5σ) of −0.20±0.35%, −0.47±0.92% and −0.06±0.17% in the head, lung and pelvis phantoms respectively, compared to the 3.5% clinical margin. Experimental HU-RSP curve have been produced on the CIRS pediatric head. Conclusion: The proposed PLSO with prior knowledge X-ray CT shows promising potential (R80 σ<1.0% over

  1. PET-CT in the evaluation of metastatic breast cancer

    International Nuclear Information System (INIS)

    Sullivan, A.M.; Fulham, M.J.

    2005-01-01

    A 44-year-old woman underwent two PET-CT scans for the evaluation of metastatic breast cancer. A radical left mastectomy with axillary dissection (1 of 43 nodes positive) followed by chemotherapy, was performed in 1998. She represented in October 2003 with a left supraclavicular fossa mass. This was confirmed to be recurrent breast cancer on FNAB. She was considered for a radical neck dissection and the surgeon requested a PET scan. Other imaging at this time included a normal bone scan and CT brain. CT neck/chest/abdomen/pelvis showed soft tissue thickening in the left lower neck. The PET-CT scan showed multiple glucose avid lesions in the sternum, mediastinum and neck lymph nodes as well as a small lesion in the proximal left femur consistent with extensive metastatic disease. Surgery was cancelled and Femara chemotherapy commenced. Femara was stopped in March 2004 and the patient began alternative therapies. In October 2004 she presented to her surgeon with new back and chest pain. CT of the neck/chest/abdomen/pelvis showed a soft tissue mass in the upper sternum and a lymph node at the base of the neck highly suspicious for metastatic disease. There were also 2 suspicious lung nodules and a lesion in the proximal left femur reported as an osteoid osteoma. Wholebody PET-CT scans were performed on a Siemens LSO Biograph, 60mins after the injection of 350Mbq of Fl 8-Fag, with arms at the patient's side and head in the field-of-view. On both occasions the patient had to pay for the scan. On the 2004 PET-CT scan, the CT brain revealed multiple hyperdense lesions consistent with hemorrhagic metastases. In addition, there were innumerable glucose avid foci involving viscera, nodes and skeleton consistent with disseminated disease. Our case illustrates: (i) the value of PET in the management of metastatic breast cancer; (ii) the improved accuracy of PET-CT in delineating sites of disease; (iii) the issues of head movement in PET-CT and. (iv) the problem with lack of

  2. Three-dimensional breast image reconstruction from a limited number of views

    Science.gov (United States)

    McCauley, Thomas G.; Stewart, Alexander X.; Stanton, Martin J.; Wu, Tao; Phillips, Walter C.

    2000-04-01

    Typically in three-dimensional (3D) computed tomography (CT) imaging, hundreds or thousands of x-ray projection images are recorded. The image-collection time and patient dose required rule out conventional CT as a tool for screening mammography. We have developed a CT method that overcomes these limitations by using (1) a novel image collection geometry, (2) new digital electronic x-ray detector technology, and (3) modern image reconstruction procedures. The method, which we call Computed Planar Mammography (CPM), is made possible by the full-field, low-noise, high-resolution CCD-based detector design that we have previously developed. With this method, we need to record only a limited number (10 - 50) of low-dose x- ray images of the breast. The resulting 3D full breast image has a resolution in two orientations equal to the full detector resolution (47 microns), and a lower, variable resolution (0.5 - 10 mm) in the third orientation. This 3D reconstructed image can then be viewed as a series of cross- sectional layers, or planes, each at the full detector resolution. Features due to overlapping tissue, which could not be differentiated in a conventional mammogram, are separated into layers at different depths. We demonstrate the features and capabilities of this method by presenting reconstructed images of phantoms and mastectomy specimens. Finally, we discuss outstanding issues related to the further development of this procedure, as well as considerations for its clinical implementation.

  3. Water equivalence of NIPAM based polymer gel dosimeters with enhanced sensitivity for x-ray CT

    Science.gov (United States)

    Gorjiara, Tina; Hill, Robin; Bosi, Stephen; Kuncic, Zdenka; Baldock, Clive

    2013-10-01

    Two new formulations of N-isopropylacrylamide (NIPAM) based three dimensional (3D) gel dosimeters have recently been developed with improved sensitivity to x-ray CT readout, one without any co-solvent and the other one with isopropanol co-solvent. The water equivalence of the NIPAM gel dosimeters was investigated using different methods to calculate their radiological properties including: density, electron density, number of electrons per grams, effective atomic number, photon interaction probabilities, mass attenuation and energy absorption coefficients, electron collisional, radiative and total mass stopping powers and electron mass scattering power. Monte Carlo modelling was also used to compare the dose response of these gel dosimeters with water for kilovoltage and megavoltage x-ray beams and for megavoltage electron beams. We found that the density and electron density of the co-solvent free gel dosimeter are more water equivalent with less than a 2.6% difference compared to a 5.7% difference for the isopropanol gel dosimeter. Both the co-solvent free and isopropanol solvent gel dosimeters have lower effective atomic numbers than water, differing by 2.2% and 6.5%, respectively. As a result, their photoelectric absorption interaction probabilities are up to 6% and 19% different from water, respectively. Compton scattering and pair production interaction probabilities of NIPAM gel with isopropanol differ by up to 10% from water while for the co-solvent free gel, the differences are 3%. Mass attenuation and energy absorption coefficients of the co-solvent free gel dosimeter and the isopropanol gel dosimeter are up to 7% and 19% lower than water, respectively. Collisional and total mass stopping powers of both gel dosimeters differ by less than 2% from those of water. The dose response of the co-solvent free gel dosimeter is water equivalent (with x-ray beams over the energy range 180 keV-18 MV, both gel dosimeters have less than 2% discrepancy with water. For

  4. Three-dimensional analysis of rodent paranasal sinus cavities from X-ray computed tomography (CT) scans

    Science.gov (United States)

    Phillips, Jonathan E.; Ji, Lunan; Rivelli, Maria A.; Chapman, Richard W.; Corboz, Michel R.

    2009-01-01

    Continuous isometric microfocal X-ray computed tomography (CT) scans were acquired from an AKR/J mouse, Brown-Norway rat, and Hartley guinea pig. The anatomy and volume of the paranasal sinus cavities were defined from 2-dimensional (2-D) and 3-dimensional (3-D) CT images. Realistic 3-D images were reconstructed and used to determine the anterior maxillary, posterior maxillary, and ethmoid sinus cavity airspace volumes (mouse: 0.6, 0.7, and 0.7 mm3, rat: 8.6, 7.7, and 7.0 mm3, guinea pig: 63.5, 46.6 mm3, and no ethmoid cavity, respectively). The mouse paranasal sinus cavities are similar to the corresponding rat cavities, with a reduction in size, while the corresponding maxillary sinus cavities in the guinea pig are different in size, location, and architecture. Also, the ethmoid sinus cavity is connected by a common drainage pathway to the posterior maxillary sinus in mouse and rat while a similar ethmoid sinus was not present in the guinea pig. We conclude that paranasal sinus cavity airspace opacity (2-D) or volume (3-D) determined by micro-CT scanning may be used to conduct longitudinal studies on the patency of the maxillary sinus cavities of rodents. This represents a potentially useful endpoint for developing and testing drugs in a small animal model of sinusitis. PMID:19794893

  5. In Depth Analyses of LEDs by a Combination of X-ray Computed Tomography (CT) and Light Microscopy (LM) Correlated with Scanning Electron Microscopy (SEM).

    Science.gov (United States)

    Meyer, Jörg; Thomas, Christian; Tappe, Frank; Ogbazghi, Tekie

    2016-06-16

    In failure analysis, device characterization and reverse engineering of light emitting diodes (LEDs), and similar electronic components of micro-characterization, plays an important role. Commonly, different techniques like X-ray computed tomography (CT), light microscopy (LM) and scanning electron microscopy (SEM) are used separately. Similarly, the results have to be treated for each technique independently. Here a comprehensive study is shown which demonstrates the potentials leveraged by linking CT, LM and SEM. In depth characterization is performed on a white emitting LED, which can be operated throughout all characterization steps. Major advantages are: planned preparation of defined cross sections, correlation of optical properties to structural and compositional information, as well as reliable identification of different functional regions. This results from the breadth of information available from identical regions of interest (ROIs): polarization contrast, bright and dark-field LM images, as well as optical images of the LED cross section in operation. This is supplemented by SEM imaging techniques and micro-analysis using energy dispersive X-ray spectroscopy.

  6. Line x-ray source for diffraction enhanced imaging in clinical and industrial applications

    Science.gov (United States)

    Wang, Xiaoqin

    Mammography is one type of imaging modalities that uses a low-dose x-ray or other radiation sources for examination of breasts. It plays a central role in early detection of breast cancers. The material similarity of tumor-cell and health cell, breast implants surgery and other factors, make the breast cancers hard to visualize and detect. Diffraction enhanced imaging (DEI), first proposed and investigated by D. Chapman is a new x-ray radiographic imaging modality using monochromatic x-rays from a synchrotron source, which produced images of thick absorbing objects that are almost completely free of scatter. It shows dramatically improved contrast over standard imaging when applied to the same phantom. The contrast is based not only on attenuation but also on the refraction and diffraction properties of the sample. This imaging method may improve image quality of mammography, other medical applications, industrial radiography for non-destructive testing and x-ray computed tomography. However, the size, and cost, of a synchrotron source limits the application of the new modality to be applicable at clinical levels. This research investigates the feasibility of a designed line x-ray source to produce intensity compatible to synchrotron sources. It is composed of a 2-cm in length tungsten filament, installed on a carbon steel filament cup (backing plate), as the cathode and a stationary oxygen-free copper anode with molybdenum coating on the front surface serves as the target. Characteristic properties of the line x-ray source were computationally studied and the prototype was experimentally investigated. SIMIION code was used to computationally study the electron trajectories emanating from the filament towards the molybdenum target. A Faraday cup on the prototype device, proof-of-principle, was used to measure the distribution of electrons on the target, which compares favorably to computational results. The intensities of characteristic x-ray for molybdenum

  7. ImaSim, a software tool for basic education of medical x-ray imaging in radiotherapy and radiology

    Science.gov (United States)

    Landry, Guillaume; deBlois, François; Verhaegen, Frank

    2013-11-01

    Introduction: X-ray imaging is an important part of medicine and plays a crucial role in radiotherapy. Education in this field is mostly limited to textbook teaching due to equipment restrictions. A novel simulation tool, ImaSim, for teaching the fundamentals of the x-ray imaging process based on ray-tracing is presented in this work. ImaSim is used interactively via a graphical user interface (GUI). Materials and methods: The software package covers the main x-ray based medical modalities: planar kilo voltage (kV), planar (portal) mega voltage (MV), fan beam computed tomography (CT) and cone beam CT (CBCT) imaging. The user can modify the photon source, object to be imaged and imaging setup with three-dimensional editors. Objects are currently obtained by combining blocks with variable shapes. The imaging of three-dimensional voxelized geometries is currently not implemented, but can be added in a later release. The program follows a ray-tracing approach, ignoring photon scatter in its current implementation. Simulations of a phantom CT scan were generated in ImaSim and were compared to measured data in terms of CT number accuracy. Spatial variations in the photon fluence and mean energy from an x-ray tube caused by the heel effect were estimated from ImaSim and Monte Carlo simulations and compared. Results: In this paper we describe ImaSim and provide two examples of its capabilities. CT numbers were found to agree within 36 Hounsfield Units (HU) for bone, which corresponds to a 2% attenuation coefficient difference. ImaSim reproduced the heel effect reasonably well when compared to Monte Carlo simulations. Discussion: An x-ray imaging simulation tool is made available for teaching and research purposes. ImaSim provides a means to facilitate the teaching of medical x-ray imaging.

  8. Development of a Radiation Dose Reporting Software for X-ray Computed Tomography (CT)

    Science.gov (United States)

    Ding, Aiping

    X-ray computed tomography (CT) has experienced tremendous technological advances in recent years and has established itself as one of the most popular diagnostic imaging tools. While CT imaging clearly plays an invaluable role in modern medicine, its rapid adoption has resulted in a dramatic increase in the average medical radiation exposure to the worldwide and United States populations. Existing software tools for CT dose estimation and reporting are mostly based on patient phantoms that contain overly simplified anatomies insufficient in meeting the current and future needs. This dissertation describes the development of an easy-to-use software platform, “VirtualDose”, as a service to estimate and report the organ dose and effective dose values for patients undergoing the CT examinations. “VirtualDose” incorporates advanced models for the adult male and female, pregnant women, and children. To cover a large portion of the ignored obese patients that frequents the radiology clinics, a new set of obese male and female phantoms are also developed and applied to study the effects of the fat tissues on the CT radiation dose. Multi-detector CT scanners (MDCT) and clinical protocols, as well as the most recent effective dose algorithms from the International Commission on Radiological Protection (ICRP) Publication 103 are adopted in “VirtualDose” to keep pace with the MDCT development and regulatory requirements. A new MDCT scanner model with both body and head bowtie filter is developed to cover both the head and body scanning modes. This model was validated through the clinical measurements. A comprehensive slice-by-slice database is established by deriving the data from a larger number of single axial scans simulated on the patient phantoms using different CT bowtie filters, beam thicknesses, and different tube voltages in the Monte Carlo N-Particle Extended (MCNPX) code. When compared to the existing CT dose software packages, organ dose data in this

  9. The possibility of using x-ray diffraction with hair to screen for pathologic conditions such as breast cancer

    International Nuclear Information System (INIS)

    James, Veronica; Cookson, David

    2000-01-01

    Mammalian hair exhibits a complex structure on length scales ranging from a few to hundreds of Angstroms. High-quality synchrotron x-ray images have yielded new insight about the structure and packing of the intermediate keratinous filaments that represent the bulk of a hair's volume. When comparing human hair diffraction patterns from healthy individuals and breast cancer patients significant differences have been seen, raising the possibility that fiber diffraction may be useful as a screening technique for certain pathologic conditions

  10. Metastatic Breast Lesion to the Falx Detected with PET/CT

    Energy Technology Data Exchange (ETDEWEB)

    Harrison, Chester; Schuster, David M. [Emory Univ., Atlanta (United States)

    2012-06-15

    Intracranial dural metastasis is increasingly encountered in imaging. Autopsies conducted on patients with advanced metastatic disease demonstrate dural involvement in 9% of cases, with breast and prostate cancer the most common primaries. Awareness of this entity and imaging appearances is especially important in evaluating malignancies prone to dural metastasis. A 57-year-old woman with a strong family history of breast cancer initially presented after self-detection of a right breast lump. Subsequent mammogram and biopsies yielded a diagnosis of right infiltrating ductal carcinoma with a positive lymph node as well as left invasive lobular carcinoma. Initial staging PET-CT (not shown) at the time of diagnosis demonstrated no abnormal FDG uptake remote from the breast. Neoadjuvant chemotherapy was instituted, and a PET-CT was obtained to evaluate disease response, demonstrating an approximately 1.8 cm hypermetabolic intra-cranial mass, localized to the region of the anterior corpus callosum on axian PET (Fig. 1a), axial fused PET-CT (Fig. 1b), and sagittal fused PET-CT (Fig. 1c) with a maximum SUV of 15.9. There was associated bifrontal vasogenic edema (Fig. 1d) on the CT demonstrated on brain windows. Marked progression of disease was noted elsewhere, including hypermetabolic adenopathy and skeletal disease. A contrast-enhanced MRI of the brain was obtained demonstrating extensive T1 hypointensity, T2, and FLAIR (Fig. 2a) hyperintensity in the bilateral paramedian frontallobes representing vasogenic edema. Post-contrast imaging demonstrated three solidly enhancing masses in the areas of described vasogenic edema, one large extra-axial and two sub-centimeter parenchymal lesions. The large extra-axial and two sub-centimeter parenchymal lesions. The large extra-axial mass demonstrated homogeneous solid enhancement, in the midline anteriorly centered on the falx, just superior to the anterior corpus callosum. This measured 1.7cm transverse x 3.1cm AP x 2.4cm

  11. X-ray differential phase-contrast tomographic reconstruction with a phase line integral retrieval filter

    International Nuclear Information System (INIS)

    Fu, Jian; Hu, Xinhua; Li, Chen

    2015-01-01

    We report an alternative reconstruction technique for x-ray differential phase-contrast computed tomography (DPC-CT). This approach is based on a new phase line integral projection retrieval filter, which is rooted in the derivative property of the Fourier transform and counteracts the differential nature of the DPC-CT projections. It first retrieves the phase line integral from the DPC-CT projections. Then the standard filtered back-projection (FBP) algorithms popular in x-ray absorption-contrast CT are directly applied to the retrieved phase line integrals to reconstruct the DPC-CT images. Compared with the conventional DPC-CT reconstruction algorithms, the proposed method removes the Hilbert imaginary filter and allows for the direct use of absorption-contrast FBP algorithms. Consequently, FBP-oriented image processing techniques and reconstruction acceleration softwares that have already been successfully used in absorption-contrast CT can be directly adopted to improve the DPC-CT image quality and speed up the reconstruction

  12. Absorption and Phase Contrast X-Ray Imaging in Paleontology Using Laboratory and Synchrotron Sources

    Energy Technology Data Exchange (ETDEWEB)

    Bidola, Pidassa; Stockmar, Marco; Achterhold, Klaus; Pfeiffer, Franz; Pacheco, Mirian L.A.F.; Soriano, Carmen; Beckmann, Felix; Herzen, Julia

    2015-10-01

    X-ray micro-computed tomography (CT) is commonly used for imaging of samples in biomedical or materials science research. Owing to the ability to visualize a sample in a nondestructive way, X-ray CT is perfectly suited to inspect fossilized specimens, which are mostly unique or rare. In certain regions of the world where important sedimentation events occurred in the Precambrian geological time, several fossilized animals are studied to understand questions related to their origin, environment, and life evolution. This article demonstrates the advantages of applying absorption and phase-contrast CT on the enigmatic fossil Corumbella werneri, one of the oldest known animals capable of building hard parts, originally discovered in Corumba (Brazil). Different tomographic setups were tested to visualize the fossilized inner structures: a commercial laboratory-based CT device, two synchrotron-based imaging setups using conventional absorption and propagation-based phase contrast, and a commercial X-ray microscope with a lens-coupled detector system, dedicated for radiography and tomography. Based on our results we discuss the strengths and weaknesses of the different imaging setups for paleontological studies.

  13. Fully automated deformable registration of breast DCE-MRI and PET/CT

    Science.gov (United States)

    Dmitriev, I. D.; Loo, C. E.; Vogel, W. V.; Pengel, K. E.; Gilhuijs, K. G. A.

    2013-02-01

    Accurate characterization of breast tumors is important for the appropriate selection of therapy and monitoring of the response. For this purpose breast imaging and tissue biopsy are important aspects. In this study, a fully automated method for deformable registration of DCE-MRI and PET/CT of the breast is presented. The registration is performed using the CT component of the PET/CT and the pre-contrast T1-weighted non-fat suppressed MRI. Comparable patient setup protocols were used during the MRI and PET examinations in order to avoid having to make assumptions of biomedical properties of the breast during and after the application of chemotherapy. The registration uses a multi-resolution approach to speed up the process and to minimize the probability of converging to local minima. The validation was performed on 140 breasts (70 patients). From a total number of registration cases, 94.2% of the breasts were aligned within 4.0 mm accuracy (1 PET voxel). Fused information may be beneficial to obtain representative biopsy samples, which in turn will benefit the treatment of the patient.

  14. A feasibility study of X-ray phase-contrast mammographic tomography at the Imaging and Medical beamline of the Australian Synchrotron.

    Science.gov (United States)

    Nesterets, Yakov I; Gureyev, Timur E; Mayo, Sheridan C; Stevenson, Andrew W; Thompson, Darren; Brown, Jeremy M C; Kitchen, Marcus J; Pavlov, Konstantin M; Lockie, Darren; Brun, Francesco; Tromba, Giuliana

    2015-11-01

    Results are presented of a recent experiment at the Imaging and Medical beamline of the Australian Synchrotron intended to contribute to the implementation of low-dose high-sensitivity three-dimensional mammographic phase-contrast imaging, initially at synchrotrons and subsequently in hospitals and medical imaging clinics. The effect of such imaging parameters as X-ray energy, source size, detector resolution, sample-to-detector distance, scanning and data processing strategies in the case of propagation-based phase-contrast computed tomography (CT) have been tested, quantified, evaluated and optimized using a plastic phantom simulating relevant breast-tissue characteristics. Analysis of the data collected using a Hamamatsu CMOS Flat Panel Sensor, with a pixel size of 100 µm, revealed the presence of propagation-based phase contrast and demonstrated significant improvement of the quality of phase-contrast CT imaging compared with conventional (absorption-based) CT, at medically acceptable radiation doses.

  15. Synchrotron nanoscopy imaging study of scalp hair in breast cancer patients and healthy individuals: Difference in medulla loss and cortical membrane enhancements.

    Science.gov (United States)

    Han, Sung-Mi; Chikawa, Jun-Ichi; Jeon, Jae-Kun; Hwang, Min-Young; Lim, Jun; Jeong, Young-Ju; Park, Sung-Hwan; Kim, Hong-Tae; Jheon, Sanghoon; Kim, Jong-Ki

    2016-01-01

    Nanoscopic synchrotron X-ray imaging was performed on scalp hair samples of patients with breast cancer and healthy individuals to investigate any structural differences as diagnostic tool. Hair strands were divided into 2-3 segments along the strands from root to tip, followed by imaging either in projection or in CT scanning with a monochromatic 6.78-keV X-ray using zone-plate optics with a resolving power of 60 nm. All the examined cancer hairs exhibited medulla loss with cancer stage-dependent pattern; complete loss, discontinuous or trace along the strands. In contrast, medullas were well retained without complete loss in the healthy hair. In the CT-scanned axial images, the cortical spindle compartments had no contrast in the healthy hair, but appeared hypointense in contrast to the surrounding hyperintense cortical membrane complex in the cancer hair. In conclusion, observation of medulla loss and cortical membrane enhancements in the hair strands of breast cancer patients demonstrated structural variations in the cancer hair, providing a new platform for further synchrotron X-ray imaging study of screening breast cancer patients. © 2015 Wiley Periodicals, Inc.

  16. Correlation between PET/CT results and histological and immunohistochemical findings in breast carcinomas

    International Nuclear Information System (INIS)

    Bitencourt, Almir Galvao Vieira; Lima, Eduardo Nobrega Pereira; Chojniak, Rubens; Marques, Elvira Ferreira; Souza, Juliana Alves de; Graziano, Luciana; Andrade, Wesley Pereira; Osorio, Cynthia Aparecida Bueno de Toledo

    2014-01-01

    Objective: to correlate the results of 18 F-fluoro-2-deoxy-D-glucose ( 18 F-FDG) positron emission tomography/computed tomography (PET/CT) performed with a specific protocol for assessment of breasts with histological/immunohistochemical findings in breast carcinoma patients. Materials and methods: cross-sectional study with prospective data collection, where patients with biopsy-confirmed breast carcinomas were studied. The patients underwent PET/CT examination in prone position, with a specific protocol for assessment of breasts. PET/CT findings were compared with histological and immunohistochemical data. Results: the authors identified 59 malignant breast lesions in 50 patients. The maximum diameter of the lesions ranged from 6 to 80 mm (mean: 32.2 mm). Invasive ductal carcinoma was the most common histological type (n = 47; 79.7%). At PET/CT, 53 (89.8%) of the lesions demonstrated anomalous concentrations of 18 F-FDG, with maximum SUV ranging from 0.8 to 23.1 (mean: 5.5). A statistically significant association was observed between higher values of maximum SUV and histological type, histological grade, molecular subtype, tumor diameter, mitotic index and Ki-67 expression. Conclusion: PET/CT performed with specific protocol for assessment of breasts has demonstrated good sensitivity and was associated with relevant histological/immunohistochemical factors related to aggressiveness and prognosis of breast carcinomas. (author)

  17. X-ray body scanner for computerised tomography

    International Nuclear Information System (INIS)

    1977-01-01

    An X-ray source is described whose source is collimated into a thin fan-shaped beam. The detector means is spaced from the sources and both are mounted for scanning and orbiting jointly about a body in a partial or complete resolution. The X-ray intensities thus obtained provide data for reconstructing an image. The detector and source combination and the body are moved relative to one another in an axial direction to enable scanning of the body layers in sequence. In one embodiment the X-ray source is pulsed as it scans, and in another the fan-shaped beam is on continuously and readout is done sequentially. Thus in either case a large number of intensities for each layer are obtained. A high precision encoder system is used to synchronize X-ray pulses and readouts spatially and with line frequency. Means are provided for storing the cables leading to the rotatable source, the detectors and other moveable components. An embodiment for scanning a body part such as a breast has means for conditioning and controlling the water in which the part is immersed. (C.F.)

  18. Phase-contrast X-ray computed tomography of non-formalin fixed biological objects

    Energy Technology Data Exchange (ETDEWEB)

    Takeda, Tohoru E-mail: ttakeda@md.tsukuba.ac.jp; Momose, Atsushi; Wu, Jin; Zeniya, Tsutomu; Yu Quanwen; Thet Thet Lwin; Itai, Yuji

    2001-07-21

    Using a monolithic X-ray interferometer having the view size of 25 mmx25 mm, phase-contrast X-ray CT (PCCT) was performed for non-formalin fixed livers of two normal rats and a rabbit transplanted with VX-2 cancer. PCCT images of liver and cancer lesions resembled well those obtained by formalin fixed samples.

  19. Developments in contact X-ray microscopy in biomedical research

    International Nuclear Information System (INIS)

    Davies, R.L.; Flores, N.A.; Pye, J.K.

    1985-01-01

    Contact X-ray microscopy (microradiography) is a method of studying the microstructure of biological tissue. These techniques have been used to study the historadiological details of human breast tissue and sections of human ear ossicles. X-ray microscopy can also be used to demonstrate variations in structural densities seen in histological specimens including the detection of microcalcification. A modification of existing apparatus is described which has resulted in improved image-contrast and detail. The ability of X-rays to penetrate relatively thick sections of tissue makes it an ideal method for studying the morphology of biological structures, particularly in calcified tissue. The tissues may be further examined by conventional histology, elemental analysis, etc. The technique has a complementary role to alternative methods of tissue microscopy. (author)

  20. Morphological Characterisation of Unstained and Intact Tissue Micro-architecture by X-ray Computed Micro- and Nano-Tomography

    Science.gov (United States)

    Walton, Lucy A.; Bradley, Robert S.; Withers, Philip J.; Newton, Victoria L.; Watson, Rachel E. B.; Austin, Clare; Sherratt, Michael J.

    2015-05-01

    Characterisation and quantification of tissue structures is limited by sectioning-induced artefacts and by the difficulties of visualising and segmenting 3D volumes. Here we demonstrate that, even in the absence of X-ray contrast agents, X-ray computed microtomography (microCT) and nanotomography (nanoCT) can circumvent these problems by rapidly resolving compositionally discrete 3D tissue regions (such as the collagen-rich adventitia and elastin-rich lamellae in intact rat arteries) which in turn can be segmented due to their different X-ray opacities and morphologies. We then establish, using X-ray tomograms of both unpressurised and pressurised arteries that intra-luminal pressure not only increases lumen cross-sectional area and straightens medial elastic lamellae but also induces profound remodelling of the adventitial layer. Finally we apply microCT to another human organ (skin) to visualise the cell-rich epidermis and extracellular matrix-rich dermis and to show that conventional histological and immunohistochemical staining protocols are compatible with prior X-ray exposure. As a consequence we suggest that microCT could be combined with optical microscopy to characterise the 3D structure and composition of archival paraffin embedded biological materials and of mechanically stressed dynamic tissues such as the heart, lungs and tendons.